Emerson Model 385 Retractable pH/ORP Sensor Instruction Manual
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Instruction Manual
PN 51-385/rev.D
September 2009
Retractable pH/ORP Sensor
Model 385
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures, and tests its
products to meet many national and international standards. Because these instruments are sophisticated technical products, you must properly install, use, and maintain
them to ensure they continue to operate within their normal
specifications. The following instructions must be adhered
to and integrated into your safety program when installing,
using, and maintaining Rosemount Analytical products.
Failure to follow the proper instructions may cause any one
of the following situations to occur: Loss of life; personal
injury; property damage; damage to this instrument; and
warranty invalidation.
• Read all instructions prior to installing, operating, and
servicing the product. If this Instruction Manual is not the
correct manual, telephone 1-800-654-7768 and the
requested manual will be provided. Save this Instruction
Manual for future reference.
• If you do not understand any of the instructions, contact
your Rosemount representative for clarification.
• Follow all warnings, cautions, and instructions marked
on and supplied with the product.
DANGER
HAZARDOUS AREA INSTALLATION
This sensor is not Intrinsically Safe. or
Explosion Proof. Installations near flammable
liquids or in hazardous area locations must
be carefully evaluated by qualified on site
safety personnel.
To secure and maintain an intrinsically safe
installation, an appropriate transmitter/
safety barrier/sensor combination must be
used. The installation system must be in
accordance with the governing approval
agency (FM, CSA or BASEEFA/CENELEC)
hazardous area classification requirements.
Consult your analyzer/transmitter instruction
manual for details.
Proper installation, operation and servicing
of this sensor in a Hazardous Area Installation is entirely the responsibility of the user.
• Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
• Install your equipment as specified in the Installation
Instructions of the appropriate Instruction Manual and
per applicable local and national codes. Connect all
products to the proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel
to install, operate, update, program, and maintain the
product.
• When replacement parts are required, ensure that qualified people use replacement parts specified by
Rosemount. Unauthorized parts and procedures can
affect the product’s performance and place the safe
operation of your process at risk. Look alike substitutions
may result in fire, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is
being performed by qualified persons, to prevent electrical shock and personal injury.
Emerson Process Management
Rosemount Analytical Inc.
2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 757-8500
Fax: (949) 474-7250
http://www.RAuniloc.com
© Rosemount Analytical Inc. 2001
WARNING
RETRACTABLE SENSORS
Retractable sensors must not be
inserted nor retracted when process
pressures are in excess of 64 psig
(442kPa).
CAUTION
SENSOR/PROCESS
APPLICATION COMPATIBILITY
The wetted sensor materials may not
be compatible with process composition and operating conditions.
Application compatibility is entirely the
responsibility of the user.
About This Document
This manual contains instructions for installation and operation of the Model 385
Retractable pH Sensor.
The following list provides notes concerning all revisions of this document.
Rev. Level
Date
Notes
A
8/98
This is the initial release of the product manual. The manual
has been reformatted to reflect the Emerson documentation
style and updated to reflect any changes in the product offering.
B
7/02
Updated drawing on page 5.
9/09
Update last page with division name and new DNV logo 2008.
C
D
Emerson Process Management
Liquid Division
2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 757-8500
Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2003
MODEL 385
TABLE OF CONTENTS
MODEL 385
RETRACTABLE pH/ORP SENSOR
TABLE OF CONTENTS
Section
Title
Page
1.0
DESCRIPTION AND SPECIFICATIONS .............................................
1
1.1
Features and Applications ....................................................................
1
1.2
Performance Specification....................................................................
1
1.3
Physical Specification...........................................................................
1
1.4
Ordering Information ............................................................................
2
2.0
INSTALLATION....................................................................................
3
2.1
Unpacking and Inspection ....................................................................
3
2.2
Mechanical Installation .........................................................................
3
2.3
Electrical Installation.............................................................................
4
3.0
START UP AND CALIBRATION..........................................................
8
3.1
Start Up ................................................................................................
8
3.2
Calibration Using Buffer Solutions or Grab Samples ...........................
8
3.3
Model 385 ORP....................................................................................
8
4.0
MAINTENANCE...................................................................................
9
4.1
Sensor Removal...................................................................................
9
4.2
Sensor Maintenance ............................................................................
9
4.3
Sensor Tube Replacement...................................................................
11
4.4
Model 385 ORP....................................................................................
12
5.0
TROUBLESHOOTING.........................................................................
14
6.0
RETURN OF MATERIAL .....................................................................
16
i
MODEL 385
TABLE OF CONTENTS
TABLE OF CONTENTS CONT’D
LIST OF FIGURES
Number
Title
Page
2-1
Typical Mounting Details.......................................................................
4
2-2
Dimensional Drawing............................................................................
5
2-3
Wiring Details for Options 50, 51, 54 ....................................................
6
2-4
Wiring Details for Option 53..................................................................
7
4-1
Preamplifier Check ...............................................................................
10
4-2
Sensor Tube Replacement ...................................................................
11
4-3
Male Connector Tightening Diagram ....................................................
13
LIST OF TABLES
Number
Title
Page
3-1
ORP of Saturated Quinhydrone Solution (In Millivolts).........................
7
4-1
Temperature vs Resistance of Auto T.C. Elements................................
10
4-2
R0 and R1 Values for Temperature Compensation Elements ..............
10
5-1
Troubleshooting ....................................................................................
14
5-2
Model 385 Replacement Parts .............................................................
15
ii
MODEL 385
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
• RETRACTABLE SENSOR can be easily removed and replaced under pressure without
process shutdown.
• CHEMICALLY RUGGED “TEFZEL” 1 BODY with Titanium tube provides maximum
resistance and is completely sealed to eliminate sensor leakage.
• LONG-LIFE, TRIPLE JUNCTION REFERENCE ELECTRODE provides longer service life in
applications where poisoning ions are present.
• INTEGRAL PREAMPLIFIER provides noise free long distance transmission of the high
impedance pH signal.
• WEATHERPROOF JUNCTION BOX permits installation in most environments.
1.1 FEATURES AND APPLICATIONS
The Rosemount Analytical Model 385 Retractable
pH/ORP Sensor is designed to meet a wide variety of
industrial applications. It has been specially designed
for applications where the process cannot be shut
down and a separate sample stream is difficult or
impossible to provide. The Model 385’s body is constructed of standard Titanium tube and Tefzel allowing
the sensor to be used in especially harsh applications
such as bleach lines in paper mills.
The Model 385 Retractable pH/ORP Sensor is
designed for use with a standard 1-1/2 in. ball valve.
This makes it ideal for replacing or updating any manufacturer's existing retractable pH/ORP sensor without
process shut-down. Initial installations can even be
accomplished without process shut-down by hot tapping a 1-1/2 in. ball valve at the desired point of measurement.
Sensor removal is accomplished by retracting the sensor until it reaches the built-in retraction stop collar (no
cables or chains are required for retraction). The ball
valve may then be closed, and the sensor removed.
The entire process is accomplished with almost no loss
of process fluid and without depressurizing the line.
Once the sensor is removed from the ball valve it can
easily be buffer checked or its disposable sensor tube
replaced.
sensor. This preamplifier provides noise free, long distance transmission of the pH/ORP signal to a remote
analyzer, such as the Rosemount Analytical Models 54
pH/ORP, 1054A pH/ORP, 3081 pH/ORP and many others. (Refer to their respective Product Data Sheet for
further information.)
1.2 PERFORMANCE SPECIFICATIONS
pH Range: ACCUGLASS™ GPLR 0-13pH, HIpH 0-14pH
Repeatability: ± 0.05 pH
Temperature Compensation: 0 to 100° C (212° F)
Maximum Process Pressure and Temperature:
100 psig (790 kPa abs) at 100° C
Maximum Pressure at Retraction and Insertion:
64 psig (542 kPa abs)
1.3 PHYSICAL SPECIFICATIONS
Junction Box: Weatherproof
Process Connection: 1-1/2 in. FNPT ball valve
1 in. MNPT with only male connector
Wetted Materials: Tefzel, titanium, Viton1, Teflon1, a
316 stainless steel compression fitting with EPDM,
glass, and ceramic
The Model 385 pH sensor comes standard with a
triple junction gel filled reference cell. The peripheral
ceramic liquid junction around the base of the glass
electrode assures maximum working life even in solutions containing ammonia, chlorine and other poisoning ions.
Interconnecting Cable: P/N 9200000
The plug-in style solid state preamplifier is housed in a
weatherproof junction box as an integral part of the
1
Electrical Classification: General Purpose
Weight/Shipping: With Ball Valve: 10 lb/15 lb
(4.5 kg/6.8 kg)
Without Ball Valve: 5 lb/10 lb (2.25 kg/4.5 kg)
Registered trademarks of E.I. du Pont de Nemours & Co.
1
MODEL 385
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.4 ORDERING INFORMATION
The Model 385 Retractable pH/ORP Sensor: Standard features include Tefzel sensor body with triple junction
gel filled reference electrode and automatic temperature compensation housed in a titanium tube. The sensor
comes with a weatherproof junction box and an internal preamplifier.
MODEL
385
CODE
02
CODE
RETRACTABLE pH/ORP SENSOR
TUBE MATERIAL (Required Selection)
Titanium
SENSOR COMPATIBILITY (Required Selection)
04
For 1181 ORP, 1060, 1023 (no T.C.)
06
For 1054/1054A/B ORP (Pt-100)
07
For 1181 pH, 1050, 1003 (3K T.C.)
08
For 1054/1054A/B pH, 2054 pH and 2081 pH (Pt-100)
CODE
COMBINATION ELECTRODE (Required Selection)
10
General Purpose pH
11
High pH
12
ORP
CODE
PREAMPLIFIER (Required Selection)
50
For use with Models 1181 pH/ORP and 1050/1060
51
For use with Models 1003/1023
52
Remote, for use with Models 1181 pH/ORP, 1050/1060, & 1003/1023, with 15 ft (4.5 m) cable
53
Remote, for use with Models 1054, 1054A, 1054B, 2054 & 2081 with 15 ft (4.5 m) cable
54
For use with Models 1054, 1054A, 1054B, 2054, 2081
CODE
BALL VALVE KIT (Optional)
20
Carbon Steel
21
316 Stainless Steel
385pH
02
08
11
54
20
EXAMPLE
NOTES:
Recommended interconnecting cable from sensor to transmitter is Belden 8722 or equivalent, available from Rosemount Analytical
as P/N 9200000. Specify length.
2
MODEL 385
SECTION 2.0
INSTALLATION
SECTION 2.0
INSTALLATION
2.1 UNPACKING AND INSPECTION. Before opening
the shipping carton, inspect the outside of the carton
for any damage. If damage is detected, contact the
carrier immediately. If there is no apparent damage,
open the carton and inspect the instrument and hardware. Make sure all the items in the packing list are
present and in good condition. Notify the factory if any
part is missing. If the instrument appears to be in satisfactory condition, proceed to Section 2.2, Mechanical
Installation.
The Model 385 is shipped in two parts. The measuring
electrode and sensor housing with junction box and the
preamplifier (or 15 ft coaxial cable for use with a remote
preamplifier). The 1-1/2 in. ball valve, 1 x 1-1/2 in.
reducing coupling and 1-1/2 in. close nipple are available as an option (P/N 23240-00 in 316 stainless steel
or P/N 23240-01 in carbon steel).
NOTE
Save the original packing cartons and
materials as most carriers require proof of
damage due to mishandling, etc. Also, if it
is necessary to return the instrument to the
factory, you must pack the instrument in
the same manner as it was received. Refer
to Section 6.0 for return instructions.
2.2.1 Installing The Sensor (Refer to Section 4.3
and Figure 4-2).
1.
If the sensor is inserted directly into the process
without a ball valve, go to Step 2. Carefully
remove the rubber boot which protects the glass
electrode and keeps the liquid junction wet. Make
sure the lubricated O-ring is in place in the groove
inside the male connector body (A). With the male
connector on the sensor, screw the male connector body (A) into the process ball valve. Insert the
sensor into the ball valve assembly until it gently
contacts the closed valve. (The molded electrode
guard will protect the glass bulb from breakage.)
2.
Thread the male connector body tightly into the
ball valve assembly or process connection. DO
NOT tighten the hex nut on the male connector
body; this would crimp the ferrule on the sensor
tube prematurely.
3.
Pull back hard on the sensor assembly, as if trying
to remove the sensor, to be certain that the sensor
cannot come free from the valve assembly and
the male connector. The built-in retraction stop
collar at the end of the sensor will butt against the
shoulder of the male connector.
CAUTION
The sensor must be captured by the valve
assembly and male connector so that it
cannot be blown free by system pressure if
mishandled during insertion or retraction.
WARNING
Glass electrode must be wetted at all times
(in storage and in line) to maximize sensor
life.
2.2 MECHANICAL INSTALLATION. The Model 385
Sensor may be installed through a 1-1/2 in. weldalet or
in a pipe tee or “Y”, as shown in Figure 2-1, when used
with a ball valve. Insert the end of the sensor to a depth
sufficient to ensure that the glass bulb is continuously
wetted by the process fluid. The Model 385 can also be
inserted directly into process without a ball valve using its
1 in. MNPT process fitting (Figure 2-2). ALLOW SUFFICIENT ROOM FOR THE SAFE MAINTENANCE OF
THE SENSOR. PERSONNEL SHOULD HAVE ROOM
FOR STABLE FOOTING DURING RETRACTION AND
RE-INSERTION OF THE SENSOR.
The sensor must be mounted within 80 degrees of
vertical with the tip pointed downward, thus keeping air
bubbles out of the pH sensitive glass bulb. Bubbles
settled in the glass bulb disrupt the electrical continuity between the pH sensitive glass and the silver/silver
chloride measuring element.
4.
When the sensor assembly is properly secured by
the ball valve assembly, the valve may be opened
and the sensor positioned into the process at the
desired depth and orientation.
5.
Tighten the hex nut of the male connector to tightly secure the sensor in place during process conditions. When tightened, the Teflon ferrule inside
the hex nut clamps the sensor tube. (See Figure
4-3).
CAUTION
Over tightening the hex nut may damage
the ferrule.
3
MODEL 385
SECTION 2.0
INSTALLATION
NOTE
A stainless steel ferrule is available if the
Teflon is inadequate. When using the
metallic ferrule, be careful and avoid over
tightening. This can damage the sensor
tube. If the male connector leaks during
insertion or retraction, replace the O-ring
in the male connector.
2.3 ELECTRICAL INSTALLATION. Make electrical
connections as shown on Figure 2-3 using the following guidelines:
1.
Pay particular attention to the analyzer or transmitter model number when following details on
the wiring diagrams to ensure that the connections are made to the proper terminals.
2.
Use a high quality four conductor shielded instrument cable such as Belden 8722. This is available
from Rosemount Analytical as Part Number
9200000.
3.
Maximum sensor to analyzer distance is:
(Integral preamplifier in sensor)
1054, 1054A pH/ORP, 1054B pH/ORP and 2054
pH - 1000 ft
1181 pH/ORP, 1050A/1060A,1003/1023 - 1 mile
with standard cable.
4.
Signal cable should be run in a dedicated conduit
and should be kept away from AC power lines.
DWG. NO.
40038503
FIGURE 2-1. Typical Mounting Details
4
REV.
D
MODEL 385
SECTION 2.0
INSTALLATION
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
DWG. NO.
40038504
REV.
F
FIGURE 2-2. Dimensional Drawing
5
MODEL 385
SECTION 2.0
INSTALLATION
DWG. NO.
40038502
FIGURE 2-3. Wiring Details for Options 50, 51, 54
6
REV.
E
MODEL 385
SECTION 2.0
INSTALLATION
DWG. NO.
40208114
REV.
A
FIGURE 2-4. Wiring Details for Option 53
7
MODEL 385
SECTION 3.0
START UP AND CALIBRATION
SECTION 3.0
START UP AND CALIBRATION
3.1 START UP. In most cases, the sensor can simply
be installed as shipped and readings with an accuracy
of ± 0.6 pH may be obtained. To obtain greater accuracy or to verify proper operation, the sensor must be
calibrated as a loop with its compatible analyzer or
transmitter. Please refer to the analyzer's or the transmitter’s respective instruction manuals for proper calibration procedures.
3.2 CALIBRATION USING BUFFER SOLUTIONS OR
GRAB SAMPLES. The sensor-analyzer/transmitter
loop may be calibrated by submersing the sensor in a
buffer solution (standard solutions of known pH values)
or in a process grab sample whose pH value may be
checked by a calibrated laboratory or portable pH meter.
3.3.1 ORP Calibration:
1.
Make a temporary electrical connection between
the sensor and the instrument.
2.
Obtain a standard solution of saturated quinhydrone (4 pH or 7 pH). This can be made quite simply by adding a few crystals of quinhydrone to
either pH 4 or pH 7 buffer. Quinhydrone is only
slightly soluble, so only a few crystals will be
required. The resulting solution is yellow in color.
(Refer to Section 4.4.1 for an alternate ORP standard solution).
3.
Immerse the sensor in the standard solution.
Allow 1-2 minutes for the ORP sensor to stabilize.
4.
Adjust the standardize control of the transmitter to
the solution value shown in Table 3-1. The resulting potentials, measured with a clean platinum
electrode and saturated KCl/AgCl reference electrode, should be within ±20 millivolts of the value
shown in Table 3-1. Solution temperature must be
noted to insure accurate interpretation of results.
The ORP value of saturated quinhydrone solution
is not stable over long periods of time; therefore,
these standards should be made up fresh each
time they are used.
5.
Remove the sensor from the buffer, rinse and
install in the process.
3.3 MODEL 385 ORP
Most industrial applications have a number of ORP
reactions occurring in sequence or simultaneously.
There can be several components that are oxidized or
reduced by the reagents that are used. Theoretically,
the ORP potential is absolute because it is the result of
the oxidation-reduction equilibrium. However, the actual measured potential is dependent on many factors,
including the condition of the surface of the ORP platinum electrode. Therefore, the sensor should be
allowed 1-2 hours to become “conditioned” to the
stream when first set-up or after being cleaned.
TABLE 3-1. ORP of Saturated Quinhydrone
Solution (In Millivolts)
Temp °C
Millivolt Potential
8
pH 4 Solution pH 7 Solution
20 25 30
20 25 30
268 264 260
94 87 80
MODEL 385
SECTION 4.0
MAINTENANCE
SECTION 4.0
MAINTENANCE
4.1 SENSOR REMOVAL. Remove the sensor from the
ball valve as follows:
WARNING
System pressure may cause the sensor to blow out with great force unless
care is taken during removal. Make
sure the following steps are adhered
to.
1.
Be certain system pressure at the sensor is below
64 psig (542 kPa) before proceeding with the
retraction. It is also recommended that the personnel wear a faceshield and have a stable footing.
2.
Push in on the sensor using the top of the J-box
and slowly loosen the hex nut (B) of the process
end male connector (A). (Refer to Figure 4-2).
3.
When the hex nut is loose enough, slowly ease
the sensor back completely until the retraction
stop collar is reached.
CAUTION
Failure to withdraw the sensor completely
may result in damage to the sensor when
the valve is closed.
4.
Close the ball valve slowly. If there is resistance it
may be hitting the sensor. Double check that the
sensor has been retracted to the retraction stop
collar.
WARNING
Before removing the sensor from the ball
valve, be absolutely certain that the ball
valve is fully closed. Leakage from the
male connector threads may indicate that
the male connector is still under pressure.
Leakage through a partially open valve
could be hazardous.
5.
The Male Connector Body (A) may now be completely unthreaded from the reducing coupling and
the sensor removed for servicing.
NOTE
With the ball valve fully closed, some residual process fluid may leak from the connectors MNPT pipe threads. This leakage
is normal and to be expected.
4.2 SENSOR MAINTENANCE.
4.2.1 Electrode Cleaning. If the electrode is coated or
dirty, clean as follows:
1.
Remove the sensor from process as previously
instructed in Section 4.1.
2.
Wipe the glass bulb with a soft, clean, lint free
cloth or tissue. If this does not remove the dirt or
coating, go to Step 3. (Detergents clean oil and
grease; and acids remove scale.)
3.
Wash the glass bulb in a strong detergent solution
and rinse it in clean water. If this does not clean
the glass bulb, go to Step 4.
CAUTION
The solution used during the following
check is an acid and should be handled
with care. Follow the directions of the acid
manufacturer. Wear the proper protective
equipment. Do not let the solution come in
contact with skin or clothing. If contact with
skin is made, immediately rinse with clean
water.
4.
Wash the glass bulb in a dilute 5% hydrochloric
acid solution and rinse with clean water. Soaking
the sensor overnight in the acid solution can
improve cleaning action.
Replace the sensor if it cannot be cleaned.
4.2.2 Automatic Temperature Compensator. The
temperature compensator element is temperature sensitive and can be checked with an ohmmeter.
Resistance increases with temperature.
The 3K element will read 3000 ohms ± 1% at 25°C
(77°F) and a Pt-100 will read 110 ohms. Resistance
varies with temperature for a 3K and Pt-100 element
and can be determined according to Table 4-1 or the
following formula:
RT=Ro [l+R1 (T-20)]
Where RT = Resistance
T = Temperature in °C
Refer to Table 4-2 for Ro and R1 values.
9
MODEL 385
SECTION 4.0
MAINTENANCE
TABLE 4-1.
Temperature vs Resistance of Auto
T.C. Elements
Temperature °C
0
10
20
25
30
40
50
60
70
80
90
100
Resistance
(Ohms) ±1%
3K
PT-100
2670
100.0
2802
103.8
2934
107.7
3000
109.6
3066
111.5
3198
115.4
3330
119.2
3462
123.1
3594
126.9
3726
130.8
3858
134.6
3990
138.5
4.2.3 Preamplifier Check (For Models 1003/1023,
1050/1060 Analyzers). To determine if the preamplifier is operable, proceed as follows:
1.
Using a BNC Shorting Cap (Refer to Figure 4-1)
or a paper clip, short the preamplifier connection.
See Table 3-2 for input voltages.
2.
With the STANDARDIZATION knob on the analyzer in a nearly vertical position, the meter should
be able to be made to read 7.0 pH and zero (0)
millivolts for ORP.
3.
If the meter/display can not be made to read 7.0 pH
or zero (0) millivolts, replace the old preamplifier with
a new one and perform the check again.
NOTE
For other model analyzers or transmitters,
refer to their respective instruction manuals for input checks, etc.
4.2.4 Preamplifier Replacement. If the preamplifier is
defective, replace as follows:
TABLE 4-2
Ro and R1 Values for Temperature
Compensation Elements
Temperature
Compensation Element
3K
PT-100
Ro
R1
2934
107.7
.0045
.00385
1.
Unscrew and remove the junction box cover.
2.
Disconnect the BNC connector and the four pin
connector from the preamplifier.
3.
Pull the preamplifier straight out.
4.
Plug in a new preamplifier, reconnect the BNC
and four pin connector, and replace the junction
box cover.
BNC SHORTING CAP
P/N 9120252
PLUG-IN STYLE PREAMP
FIGURE 4-1. Preamplifier Check
10
MODEL 385
SECTION 4.0
MAINTENANCE
4.3 SENSOR TUBE REPLACEMENT (Refer to
Figure 4-2). Replacement of a Model 385 Sensor
involves the removal and installation of two sets of
male connectors, one at the process end of the sensor,
and the other at the junction box end. Refer to
Section 4.1 for removal of the sensor.
1. The junction box with attached male connector
must be recovered from the old sensor for reuse.
Unscrew the junction box cover and set aside.
Disconnect electrical connections from printed circuit board inside junction box. Disconnect BNC
connector to preamp. Unscrew hex nut (D) from
male connector body (C). Separate junction box
from used sensor. Set aside.
2. Pry off split ferrule from sensor and set aside for
reuse. Remove hex nut (D) and set aside for
reuse. Check that the internal O-ring is in place in
the male connector body (C) attached to the junction box.
3. Remove hex nut (B) from male connector body (A)
at process end of sensor and set aside. Slide the
Teflon ferrule and the male connector off sensor in
the direction of junction box and set aside. Discard
sensor tube.
NOTE
If stainless steel ferrule was used, male
connector body (A) will have to be discard-
4.
ed with the sensor tube. Save male connector if Teflon ferrule was used. Set aside
for future use.
Discard used O-ring from male connector body
(A). Coat new O-ring with a thin film of the O-ring
lubricant provided. Position it in the machined Oring groove in place of the discarded O-ring.
CAUTION
Make sure lubricant does not contact any part
of the sensor tip particularly the glass bulb.
5. Cover the 1" MNPT pipe threads of the male connector body (A) with Teflon tape (not provided) to
protect them from galling during reinstallation.
6. Pass the wires from the new sensor through the
process end male connector (A). Make sure that
the beveled edge of the ferrule faces the process
end of the sensor. Snug the hex nut (B) to keep it
in place. Do not tighten down fully on the hex nut at
this time.
7. Pass the wires from the new sensor through the hex
nut (D), the split ferrule (from the old sensor), male
connector body (C), O-ring, and through the junction
box from the “neck” opening and out through the
space between the printed circuit board and the side
of the junction box. Butt the ferrules beveled edge
and the sensor tube against
FIGURE 4-2. Sensor Tube Replacement
11
MODEL 385
4.3 Sensor Tube Replacement (continued)
the junction male connector (C). Screw the hex nut
(D) by hand until the tube is “locked” into the male
connector body. Make sure that the male connector
body (C) is sufficiently tightened. The sensor will
“click” into place by pulling the sensor tube away
from the junction box, but will not move from side to
side or pull clear of the male connector. If the sensor tube is correctly attached to the junction box,
wrench tighten hex nut (D) on male connector body
(C). (See Figure 4-3.) Do not put the sensor tube in
a vise or use a pipe wrench to tighten the hardware
as these will damage the sensor. If sensor tube is
not correctly attached to the junction box, loosen
hex nut (D) and repeat.
8.
9.
Connect the sensor wires to the terminals on the
printed circuit board in the junction box in the manner recommended on the junction box cover, or
see Figure 2-3. Reattach the BNC connector to the
preamp. Screw on the cover of the junction box.
Insert the sensor in the process fitting. Stop it
against the closed ball valve. Slide the processend male connector down the sensor tube to mate
with the process fitting. Tighten the male connector into the process fitting.
10. Pull back hard on the sensor assembly, as if trying
to remove the sensor, to be certain that the sensor
cannot come free from the valve assembly and
male connector. The built-in retraction stop collar
at the end of the sensor will butt against the shoulder of the male connector.
11. Open ball valve and position the sensor at the
desired insertion depth and orientation. Using a
crescent or open end wrench, tighten the hex nut
(B) to secure the sensor in place.
NOTE
A stainless steel ferrule is available if the
Teflon is inadequate. When using the
metallic ferrule, be careful and avoid over
tightening. This can damage the sensor
tube.
If the male connector leaks during insertion
or retraction, replace the O-Ring in the
male connector body (A).
If the sensor is to be stored, the rubber boot
should be filled with 7 pH buffer solution
and replaced on sensor tip until ready to
use.
12
SECTION 4.0
MAINTENANCE
4.4 MODEL 385 ORP
4.4.1 Platinum Electrode Check. The platinum electrode may be checked as follows: There are two types
of standard solutions which may be used to check the
ORP electrode/transmitter system.
Type 1: One type of commonly used ORP standard
solution is the saturated quinhydrone solution. Refer to
Section 3.3.
CAUTION
The solution used during the following
check is an acid and should be handled
with care. Follow the directions of the acid
manufacturer. Wear the proper protective
equipment. If contact with skin of clothing is
made, immediately rinse with plenty of
clean water.
Type 2: A second ORP standard solution is the FerricFerrous Ammonium Sulfate Solution (PN R508-16OZ),
and it can be ordered as a spare part; otherwise, it can
be prepared from the following recipe: Dissolve 39.2
grams or reagent grade ferrous ammonium sulfate,
Fe(NH4)2 (SO4)2 • 6H2O and 48.2 grams of reagent
grade ferric ammonium sulfate, FeNH4(SO4)2 •
12H2O, in approximately 700 milliliters of water (distilled water is preferred, but tap water is acceptable).
Slowly and carefully add 56.2 milliliters of concentrated
sulfuric acid. Add sufficient water to bring the total solution volume up to 1000 ml. This standard ORP solution,
although not as simple to prepare as the quinhydrone
recipe, is much more stable, and will maintain its millivolt value for approximately one year when stored in
glass containers. This solution (ferric/ferrous ammonium sulfate) will produce a nominal ORP of 476 +20 mV
at 25°C when used with a saturated KCl/AgCl reference electrode and platinum measuring electrode.
Some tolerance in mV values is to be expected due to
the rather large liquid reference junction potentials
which can arise when measuring this strongly acidic
and concentrated solution. However, if the measuring
electrodes are kept clean and in good operating condition, consistently repeatable calibrations can be carried
out using this standard solution.
4.4.2 Cleaning Platinum Electrode. If the electrode is
never exposed to undesirable compounds, the electrode
can be restored to normal operation by simply cleaning
the platinum electrode with baking soda. Polish it by rubbing it with a damp paper towel and baking soda until a
bright, shiny appearance is attained.
MODEL 385
SECTION 4.0
MAINTENANCE
FIGURE 4-3. Male Connector Tightening Diagram
13
MODEL 385
SECTION 5.0
TROUBLESHOOTING
SECTION 5.0
TROUBLESHOOTING
TROUBLESHOOTING. In the event of malfunction, refer to Table 5-1 below. This is intended as a guide and lists
the troubles in order of probable frequency of occurrence. Do not be misled by the trouble, always look for the
cause before trying the remedy. Table 5-2 is the replacement parts list.
TABLE 5-1. Troubleshooting
Trouble
Meter reads off scale. (Display
reads overrange).
Probable Cause
Defective preamplifier.
Remedy
Check preamplifier as instructed
in Section 4.2.3 and replace
preamplifier if defective.
T.C. element shorted.
Check T.C. element as instructed in Section 4.2.2 and replace
sensor tube if defective.
Electrode not in solution or sample
or stream is not full.
Make sure sensor is in solution
that sample stream is full (see
Section 2.0 for installation
details).
Open glass electrode.
Replace sensor tube.
Reference element open - no contact.
Replace sensor tube.
Display reads between 3 and 6 pH
regardless of actual pH of solution
or sample.
Electrode cracked.
Replace sensor tube.
Meter or display indication swings
jumps widely in AUTO T.C.
Mode.
T.C. element shorted.
Check T.C. element as instructed
ed in Section 4.2.2 and replace
sensor tube if defective.
Span between buffers extremely
short in AUTO T.C. Mode.
T.C. element open.
Check T.C. element as instruct
ed in Section 4.2.2 and replace
sensor tube if defective.
Sluggish or slow meter indication
for real changes in pH level.
Electrode coated.
Clean sensor as instructed in
Sections 4.2.1 or 4.4.2.
Electrode defective.
Replace sensor tube.
Electrode coated or cracked.
Clean Sensor as instructed in
Sections 4.2.1 or 4.4.2 and,
if cracked,replace sensor tube.
Defective preamplifier.
Check preamplifier as instructed
in Section 4.2.3 and replace
preamplifier if defective.
Old glass electrode (greater than
1 year old).
Replace sensor tube.
Coated glass.
Clean Sensor as instructed in
Sections 4.2.1 or 4.4.2.
Transmitter cannot be standardized.
Transmitter short spans between
two different buffer values.
14
MODEL 385
SECTION 5.0
TROUBLESHOOTING
TABLE 5-2. Model 385 Replacement Parts
PART NUMBER
22698-00
22698-02
22698-03
23166-00
23166-01
23240-00
23240-01
23239-04
23239-05
23239-06
23239-07
23239-08
23239-09
33046-00
33054-00
9200000
9210012
9210013
9210014
9310094
9310096
9310100
9550167
DESCRIPTION
Preamp for 1003/1023 (Code 51)
Preamp for 1181pH/ORP, 1050/1060 (Code 50)
Preamp for 1054pH/ORP (Code 54)
Connector Kit, 1x1" 316 SS, with O-Ring Groove
Connector Kit, 1x1" Titanium, with O-Ring Groove
Ball valve Kit, 1-¼" 316 SS (Code 21)
Ball valve Kit, 1-¼" Carbon Steel (Code 20)
Kit, pH Sensor Tube w/3K TC, Titanium, General Purpose (Code 07-10, P/N 23288-08)3
Kit, pH Sensor Tube w/PT 100 TC, Titanium, General Purpose (Code 08-10, P/N 23288-09)3
Kit, ORP Sensor Tube w/No TC, Titanium (Code 04-12, P/N 23288-12)3
Kit, ORP Sensor Tube w/PT 100 TC, Titanium (Code 06-12, P/N 23288-13)3
Kit, pH Sensor Tube w/3K TC, Titanium, High pH (Code 07-11, P/N 23288-16)3
Kit, pH Sensor Tube w/PT 100 TC, Titanium, High pH (Code 08-11, P/N 23288-17)3
Ferrule, 1" 316SS, Split (for J-Box)
Male Connector, 1" X 1" 316SS, with O-Ring Groove, Process End
Cable, 20#Ga, 4 Cond, 1 Shielded Pair (Belden 8722)
pH Buffer, 4.01 pH, 16 oz.
pH Buffer, 6.86 pH, 16 oz.
pH Buffer, 9.18 pH, 16 oz.
Ferrule, 1" 316 SS
Hex Nut, 1" 316 SS
Ferrule, 1" Teflon (Standard)
O-Ring, 2-214 EPDM (Standard)
3 Kit includes P/N 9310100 and P/N 9550099
15
MODEL 385
SECTION 6.0
RETURN OF MATERIAL
SECTION 6.0
RETURN OF MATERIAL
6.1 GENERAL.
To expedite the repair and return of instruments, proper communication between the customer and the factory is
important. Before returning a product for repair, call 1-949-757-8500 for a Return Materials Authorization (RMA)
number.
6.2 WARRANTY REPAIR.
The following is the procedure for returning instruments still under warranty:
1.
Call Rosemount Analytical for authorization.
2.
To verify warranty, supply the factory sales order number or the original purchase order number. In the case
of individual parts or sub-assemblies, the serial number on the unit must be supplied.
3.
Carefully package the materials and enclose your “Letter of Transmittal” (see Warranty). If possible, pack the
materials in the same manner as they were received.
4.
Send the package prepaid to:
Rosemount Analytical Inc., Uniloc Division
Uniloc Division
2400 Barranca Parkway
Irvine, CA 92606
Attn: Factory Repair
RMA No. ____________
Mark the package: Returned for Repair
Model No. ____
6.3 NON-WARRANTY REPAIR.
The following is the procedure for returning for repair instruments that are no longer under warranty:
1.
Call Rosemount Analytical for authorization.
2.
Supply the purchase order number, and make sure to provide the name and telephone number of the individual to be contacted should additional information be needed.
3.
Do Steps 3 and 4 of Section 6.2.
NOTE
Consult the factory for additional information regarding service or repair.
16
WARRANTY
Seller warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufactured
or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expiration of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen
(18) months from the date of shipment by Seller, whichever period expires first. Consumables, such as glass electrodes,
membranes, liquid junctions, electrolyte, o-rings, catalytic beads, etc., and Services are warranted for a period of 90
days from the date of shipment or provision.
Products purchased by Seller from a third party for resale to Buyer ("Resale Products") shall carry only the warranty extended by
the original manufacturer. Buyer agrees that Seller has no liability for Resale Products beyond making a reasonable commercial
effort to arrange for procurement and shipping of the Resale Products.
If Buyer discovers any warranty defects and notifies Seller thereof in writing during the applicable warranty period, Seller shall, at
its option, promptly correct any errors that are found by Seller in the firmware or Services, or repair or replace F.O.B. point of manufacture that portion of the Goods or firmware found by Seller to be defective, or refund the purchase price of the defective portion of the Goods/Services.
All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuitable environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any other
cause not the fault of Seller are not covered by this limited warranty, and shall be at Buyer's expense. Seller shall not be obligated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by
an authorized Seller representative. All costs of dismantling, reinstallation and freight and the time and expenses of Seller's personnel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Seller.
Goods repaired and parts replaced during the warranty period shall be in warranty for the remainder of the original warranty period or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Seller and can be amended only
in a writing signed by an authorized representative of Seller. Except as otherwise expressly provided in the Agreement, THERE
ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESS OR IMPLIED, AS TO MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES.
RETURN OF MATERIAL
Material returned for repair, whether in or out of warranty, should be shipped prepaid to:
Emerson Process Management
Liquid Division
2400 Barranca Parkway
Irvine, CA 92606
The shipping container should be marked:
Return for Repair
Model _______________________________
The returned material should be accompanied by a letter of transmittal which should include the following information (make a
copy of the "Return of Materials Request" found on the last page of the Manual and provide the following thereon):
1.
2.
3.
4.
5.
Location type of service, and length of time of service of the device.
Description of the faulty operation of the device and the circumstances of the failure.
Name and telephone number of the person to contact if there are questions about the returned material.
Statement as to whether warranty or non-warranty service is requested.
Complete shipping instructions for return of the material.
Adherence to these procedures will expedite handling of the returned material and will prevent unnecessary additional charges
for inspection and testing to determine the problem with the device.
If the material is returned for out-of-warranty repairs, a purchase order for repairs should be enclosed.
The right people,
the right answers,
right now.
ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
http://www.raihome.com
Specifications subject to change without notice.
8
Credit Cards for U.S. Purchases Only.
Emerson Process Management
Rosemount Analytical
2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 757-8500
Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2008