Yokogawa Model ZR22G, ZR402G Separate type Zirconia Oxygen/Humidity Analyzer User's Manual
User’s
Manual
Model ZR22G, ZR402G
Separate type Zirconia
Oxygen/Humidity Analyzer
IM 11M12A01-02E
IM 11M12A01-02E
12th Edition
i
u Introduction
Thank you for purchasing the ZR22G, ZR402G Separate type Zirconia Oxygen/Humidity
Analyzer.
Please read the following respective documents before installing and using the ZR22G, ZR402G
Separate type Zirconia Oxygen/Humidity Analyzer.
The related documents are as follows.
General Specifications
Contents
Model ZR22G, ZR402G, and ZR202G
Direct In Situ Zirconia Oxygen Analyzers and High
Temperature Humidity Analyzers
Document number
Note
GS 11M12A01-01E
* the “E” in the document number is the language code.
User’s Manual
Contents
Model ZR22G, ZR402G
Separate type Zirconia Oxygen/Humidity Analyzer
Model ZR22S, ZR402G
Separate type Explosion-proof Zirconia Oxygen Analyzer
Model ZR22A, ZR202A
Heater Assembly
Model EXAxt ZR Series
HART Protocol
Model AV550G Zirconia Oxygen Analyzer Averaging
Converter
Document number
Note
IM 11M12A01-02E
(This manual)
IM 11M13A01-02E
When you use the ZR22S
Explosion-proof detector,
read this manual.
IM 11M12A01-21E
IM 11M12A01-51E
IM 11M12D01-01E
* the “E” in the document number is the language code.
An exclusive User’s Manual might be attached to the products whose suffix codes or option codes
contain the code “Z” (made to customers’ specifications). Please read it along with this manual.
The EXAxt ZR Separate type Zirconia Oxygen/Humidity Analyzer is usually the Oxygen
Analyzer which connected ZR402G converter and ZR22G detector together, but it is to the
High Temperature Humidity Analyzer when the option code “/HS (Set for Humidity Analyzer)” of
ZR402G is selected.
After that, in this manaul, the Humidity Analyzer refers to the thing which ZR22G combined with
ZR402G with “/HS”.
In addition, in this manual, the Oxygen Analyzer is mainly listed. When there are not mentions
such as “in the case of Humidity Analyzer”, it becomes same as the Oxygen Analyzer.
The EXAxt ZR Separate type Zirconia Oxygen/Humidity Analyzer has been developed for
combustion control in various industrial processes. This analyzer basically consists of a detector
and a converter. You can select between several versions, based upon your application.
Optional accessories are also available to improve measurement accuracy and automate
calibration. An optimal control system can be realized by adding appropriate options.
This instruction manual refers to almost all of the equipment related to the EXAxt ZR.
You may skip any section(s) on the equipment which is not included in your system.
Regarding the HART communication protocol, refer to IM 11M12A01-51E.
IM 11M12A01-51E has been published as “Model EXAxt ZR Series HART Protocol” .
The Integrated type (with sensor and analyzer integrated in one body) is described in
IM 11M12A01-04E.
Media No. IM 11M12A01-02E
12th Edition : Apr. 2017 (YK)
All Rights Reserved Copyright © 2000, Yokogawa Electric Corporation
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
ii
Models and descriptions in this manual are listed below.
Models and descriptions in this manual
Model
Description in this manual
Specification Installation Operation Maintenance
General use detector
○
○
○
○
High temperature detector (0.15m)
○
○
○
○
Converter
○
○
○
○
Probe protector
○
○
High temperature probe adapter
○
○
Dust protector (only for Humidity
○
○
analyzer)
Flow setting unit (for manual
○
○
○
calibration use)
Automatic Calibration unit
○
○
○
Auxiliary ejector for high
temperature use (Part No.
○
○
E7046EC, E7046EN)
Calibration gas unit case (Part No.
○
○
E7044KF)
Check valve (Part No. K9292DN,
○
○
K9292DS)
Dust filter for the detector (Part
○
○
No. K9471UA)
Dust guard protector (Part No.
○
○
K9471UC)
Standard gas unit
○
○
○
Product Name
ZR22G
ZR22G
ZR402G
ZO21R
ZO21P
ZH21B
ZA8F
ZR40H
ZO21S
CMPL
○
○
○
○
○
○
CMPL: Customer Maintenance Parts List
This manual consists of twelve chapters. Please refer to the reference chapters for installation,
operation and maintenance.
Table of Contents
Chapter
1. Overview
2. Specifications
3. Installation
4. Piping
5. Wiring
6. Components
7. Startup
8. Detailed Data Setting
9. Calibration
10. Other Functions
11. Inspection and
Maintenance
12. Troubleshooting
CMPL (parts list)
Outline
Equipment models and system
configuration examples
Standard specification, model code (or part
number), dimension drawing for each equipment
Installation method for each equipment
Examples of piping in three standard system
configurations
Wiring procedures such as “Power supply
wiring”, “output signal wiring” or others
Major parts and function are described in
this manual
Basic procedure to start operation of EXAxt
ZR. Chapter 7 enables you to operate the
equipment immediately.
Details of key operations and displays
Describes the calibration procedure
required in the course of operation.
Other functions described
How to conduct maintenance of EXAxt ZR and
procedures for replacement of deteriorated parts
This chapter describes measures to be
taken when an abnormal condition occurs.
User replaceable parts list
Relates to
Installation Operation Maintenance
B
C
B
A
B
B
A
C
A
C
A
C
C
A: Read and completely understand before operating the equipment.
B: Read before operating the equipment, and refer to it whenever necessary.
C: Recommended to read it at least once.
IM 11M12A01-02E
B
B
A
C
B
C
B
C
B
A
B
A
C
B
C
B
12th Edition : Apr. 27, 2017-00
iii
n For the safe use of this equipment
WARNING
Handle it with care. Be sure not to accidentally drop it.
Handle safely to avoid injury.
Connect the power supply cord only after confirming that the supply voltage matches the rating of
this equipment. In addition, confirm that the power is switched off when connecting power supply.
Some sample gas is dangerous to people. When removing this equipment from the process line
for maintenance or other reasons, protect yourself from potential poisoning by using a protective
mask or ventilating the area well.
CAUTION
The cell (sensor) at the tip of the detector is made of ceramic (zirconia element). Do not drop
the detector or subject it to pressure stress.
• Do NOT allow the sensor (probe tip) to make contact with anything when installing the
detector.
• Avoid any water dropping directly on the probe (sensor) of the detector when installing it.
• Check the calibration gas piping before introducing the calibration gas to ensure that there
is no leakage of the gas. If there is any leakage of the gas, the moisture drawn from the
sample gas may damage the sensor.
• The detector (especially at the tip) becomes very hot. Be sure to handle it with gloves.
n NOTICE
l Specification check
When the instrument arrives, unpack the package with care and check that the instrument
has not been damaged during transportation. In addition, please check that the specification
matches the order, and required accessories are not missing. Specifications can be
checked by the model codes on the nameplate. Refer to Chapter 2 Specifications for the list
of model codes.
l Details on operation parameters
When the EXAxt ZR Separate type Oxygen Analyzer arrives at the user site, it will operate
based on the operation parameters (initial data) set before shipping from the factory.
Ensure that the initial data is suitable for the operation conditions before conducting
analysis. Where necessary, set the instrument parameters for appropriate operation.
For details of setting data, refer to chapters 7 to 10.
When user changes the operation parameter, it is recommended to note down the changed
setting data.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
iv
u
Safety Precautions
n Safety, Protection, and Modification of the Product
• In order to protect the system controlled by the product and the product itself and ensure
safe operation, observe the safety precautions described in this user’s manual. We assume
no liability for safety if users fail to observe these instructions when operating the product.
• If this instrument is used in a manner not specified in this user’s manual, the protection
provided by this instrument may be impaired.
• If any protection or safety circuit is required for the system controlled by the product or for
the product itself, prepare it separately.
• Be sure to use the spare parts approved by Yokogawa Electric Corporation (hereafter
simply referred to as YOKOGAWA) when replacing parts or consumables.
• Modification of the product is strictly prohibited.
• The following safety symbols are used on the product as well as in this manual.
WARNING
This symbol indicates that an operator must follow the instructions laid out in this manual in order
to avoid the risks, for the human body, of injury, electric shock, or fatalities. The manual describes
what special care the operator must take to avoid such risks.
CAUTION
This symbol indicates that the operator must refer to the instructions in this manual in order to
prevent the instrument (hardware) or software from being damaged, or a system failure from
occurring.
CAUTION
This symbol gives information essential for understanding the operations and functions.
NOTE
This symbol indicates information that complements the present topic.
This symbol indicates Protective Ground Terminal.
This symbol indicates Function Ground Terminal. Do not use this terminal as the protective
ground terminal.
n Warning and Disclaimer
The product is provided on an “as is” basis. YOKOGAWA shall have neither liability nor
responsibility to any person or entity with respect to any direct or indirect loss or damage arising
from using the product or any defect of the product that YOKOGAWA can not predict in advance.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
v
n Notes on Handling User’s Manuals
• Please hand over the user’s manuals to your end users so that they can keep the user’s
manuals on hand for convenient reference.
• Please read the information thoroughly before using the product.
• The purpose of these user’s manuals is not to warrant that the product is well suited to any
particular purpose but rather to describe the functional details of the product.
• No part of the user’s manuals may be transferred or reproduced without prior written
consent from YOKOGAWA.
• YOKOGAWA reserves the right to make improvements in the user’s manuals and product at
any time, without notice or obligation.
• If you have any questions, or you find mistakes or omissions in the user’s manuals, please
contact our sales representative or your local distributor.
n Drawing Conventions
Some drawings may be partially emphasized, simplified, or omitted, for the convenience of
description.
Some screen images depicted in the user’s manual may have different display positions or
character types (e.g., the upper / lower case). Also note that some of the images contained in this
user’s manual are display examples.
In the figure listed in this manual, the example of the oxygen analyzer is shown mainly.
In the case of the humidity analyzer, unit indication may be different. Please read it appropriately.
n Product Disposal:
The instrument should be disposed of in accordance with local and national legislation/regulations.
n Trademark Acknowledgments
• All other company and product names mentioned in this user’s manual are trademarks or
registered trademarks of their respective companies.
• We do not use TM or ® mark to indicate those trademarks or registered trademarks in this
user’s manual.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
vi
CE marking products
u
n Authorized Representative in EEA
The Authorized Representative for this product in EEA is Yokogawa Europe B.V. (Euroweg 2,
3825 HD Amersfoort, The Netherlands).
n Identification Tag
This manual and the identification tag attached on packing box are essential parts of the product.
Keep them together in a safe place for future reference.
n Users
This product is designed to be used by a person with specialized knowledge.
n How to dispose the batteries:
This is an explanation about the new EU Battery Directive (DIRECTIVE 2006/66/EC). This
directive is only valid in the EU.
Batteries are included in this product. Batteries incorporated into this product cannot be
removed by yourself. Dispose them together with this product.
When you dispose this product in the EU, contact your local Yokogawa Europe B.V.office.
Do not dispose them as domestic household waste.
Battery type: Manganese dioxide lithium battery
Notice:
The symbol (see above) means they shall be sorted out and collected as ordained in
ANNEX II in DIRECTIVE 2006/66/EC.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
Toc-1
Model ZR22G, ZR402G
Separate type Zirconia
Oxygen/Humidity Analyzer
IM 11M12A01-02E 12th Edition
CONTENTS
u Introduction.....................................................................................................i
u
Safety Precautions.......................................................................................iv
u
CE marking products...................................................................................vi
1. Overview..................................................................................................... 1-1
1.1
1.2
2.
< EXAxt ZR > System Configuration................................................................ 1-1
1.1.1
System 1............................................................................................. 1-1
1.1.2
System 2............................................................................................. 1-2
1.1.3
System 3............................................................................................. 1-3
< EXAxt ZR > System Components................................................................. 1-4
1.2.1
System Components.......................................................................... 1-4
1.2.2
Detectors and Accessories................................................................. 1-4
Specifications............................................................................................ 2-1
2.1
General Specifications...................................................................................... 2-1
2.2
General use Separate type Detector and Related Equipment...................... 2-3
2.3
2.2.1
ZR22G General use Separate type Detector..................................... 2-3
2.2.2
ZO21R Probe Protector...................................................................... 2-7
2.2.3
ZH21B Dust Protector......................................................................... 2-8
Separate type Detector for High Temperature and Related Equipment...... 2-9
2.3.1
ZR22G (0.15m) Separate type Detector for High Temperature......... 2-9
2.3.2
ZO21P High Temperature Probe Adapter ........................................ 2-10
2.4
ZR402G Separate type Converter.................................................................. 2-11
2.5
ZA8F Flow Setting Unit and
ZR40H Automatic Calibration Unit................................................................. 2-17
2.5.1
ZA8F Flow Setting Unit..................................................................... 2-17
2.5.2
ZR40H Automatic Calibration Unit.................................................... 2-19
2.6
ZO21S Standard Gas Unit............................................................................... 2-21
2.7
Other Equipments............................................................................................ 2-22
2.7.1
Dust Filter for the Detector (K9471UA)............................................. 2-22
2.7.2
Dust Guard Protector (K9471UC)..................................................... 2-23
2.7.3
Ejector Assembly for High Temperature (E7046EC, E7046EN)...... 2-23
2.7.4
Stop Valve (L9852CB, G7016XH).................................................... 2-25
2.7.5
Check Valve (K9292DN, K9292DS)................................................. 2-26
2.7.6
Air Set................................................................................................ 2-27
2.7.7
Zero Gas Cylinder (G7001ZC)......................................................... 2-28
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
Toc-2
3.
2.7.8
Cylinder Pressure Reducing Valve (G7013XF, G7014XF).................2-28
2.7.9
Case Assembly for Calibration Gas Cylinder (E7044KF)................. 2-29
2.7.10
ZR22A Heater Assembly.................................................................. 2-30
Installation.................................................................................................. 3-1
3.1
3.2
Installation of General-purpose Detector........................................................ 3-1
3.1.1
Probe Insertion Hole........................................................................... 3-1
3.1.2
Installation of the Detector.................................................................. 3-2
3.1.3
Installation of the Dust Filter (K9471UA), Dust Guard Protector
(K9471UC), Probe Protector ZO21R.................................................. 3-2
3.1.4
Installation of ZH21B Dust Protector.................................................. 3-4
Installation of High Temperature Detector (ZR22G-015)............................... 3-5
3.2.1
Usage of the High Temperature Probe Adapter (ZO21P-H)............... 3-5
3.2.2
Probe Insertion Hole........................................................................... 3-6
3.2.3
Mounting of the High Temperature Detector....................................... 3-7
3.3
Installation of the ZR402G Converter.............................................................. 3-8
3.4
Installation of ZA8F Flow Setting Unit........................................................... 3-10
3.5
Installation of ZR40H Automatic Calibration Unit........................................ 3-11
3.6
Installation of the Case Assembly (E7044KF)................................................3-12
3.7
Insulation Resistance Test.............................................................................. 3-13
4. Piping.......................................................................................................... 4-1
4.1
4.2
Piping for System 1............................................................................................ 4-1
4.1.1
Parts Required for Piping in System 1................................................ 4-2
4.1.2
Connection to the Calibration Gas Inlet.............................................. 4-3
4.1.3
Connection to the Reference Gas Inlet.............................................. 4-3
4.1.4
Piping to the High Temperature Probe Adapter.................................. 4-3
Piping for System 2............................................................................................ 4-5
4.2.1
Piping Parts for System 2................................................................... 4-6
4.2.2
Piping for the Calibration Gas............................................................. 4-6
4.2.3
Piping for the Reference Gas............................................................. 4-7
4.2.4
Piping to the High Temperature Probe Adapter.................................. 4-7
4.3
Piping for System 3............................................................................................ 4-7
4.4
Piping for the Detector with Pressure Compensation................................... 4-9
4.4.1
Piping Parts for a System using Detector with Pressure Compensation.
4-11
4.4.2
Piping for the Calibration Gas........................................................... 4-11
4.4.3
Piping for the Reference Gas........................................................... 4-11
5. Wiring.......................................................................................................... 5-1
5.1
General................................................................................................................ 5-1
5.1.1
Terminals for the External Wiring in the Converter............................. 5-2
5.1.2 Wiring.................................................................................................. 5-3
5.1.3
Mounting of Cable Gland.................................................................... 5-4
5.2 Wiring.................................................................................................................. 5-5
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
Toc-3
6.
5.2.1
Connection to Converter..................................................................... 5-5
5.2.2
Connection to Detector....................................................................... 5-5
5.2.3
Power and Grounding Wiring............................................................. 5-6
5.2.4
Wiring for Power to Detector Heater................................................... 5-7
5.2.5
Wiring for Detector Output.................................................................. 5-7
5.2.6
Wiring for Analog Output..................................................................... 5-9
5.2.7
Contact Output Wiring........................................................................ 5-9
5.2.8
Contact Input Wiring......................................................................... 5-10
5.2.9
Wiring for ZR40H Automatic Calibration Unit................................... 5-11
5.2.10
Temperature Input Wiring................................................................. 5-12
Components.............................................................................................. 6-1
6.1
ZR22G Detector.................................................................................................. 6-1
6.1.1
General-purpose Detector (except for ZR22G-015).......................... 6-1
6.1.2
High Temperature Detector (ZR22G-015).......................................... 6-2
6.2
ZR402G Converter............................................................................................. 6-3
6.3
Touchpanel Switch Operations........................................................................ 6-4
6.4
6.3.1
Basic Panel and Switch...................................................................... 6-4
6.3.2
Display Configuration.......................................................................... 6-4
6.3.3
Display Functions................................................................................ 6-5
6.3.4
Entering Numeric and Text Data......................................................... 6-6
ZA8F Flow Setting Unit, ZR40H Automatic Calibration Unit......................... 6-8
7. Startup........................................................................................................ 7-1
7.1
Startup Procedure.............................................................................................. 7-1
7.2
Checking Piping and Wiring Connections...................................................... 7-1
7.3
Checking Valve Setup........................................................................................ 7-2
7.4
Supplying Power to the Converter................................................................... 7-2
7.5
Confirmation of Converter Type Setting......................................................... 7-3
7.6
Confirmation of Detector Type Setting............................................................ 7-4
7.7
Selection of Sample Gas................................................................................... 7-4
7.8
Output Range Setting........................................................................................ 7-4
7.9
7.8.1
Oxygen Analyzer - Minimum Current (4 mA) and Maximum
Current (20 mA) Settings.................................................................... 7-4
7.8.2
Humidity Analyzer - Analog Output Setting......................................... 7-5
7.8.3
Humidity Analyzer - Minimum Current (4 mA) and Maximum
Current (20 mA) Settings.................................................................... 7-5
Setting Display Item........................................................................................... 7-6
7.9.1
Oxygen Analyzer - Setting Display Item............................................. 7-6
7.9.2
Humidity Analyzer - Setting Display Item............................................ 7-8
7.10
Checking Current Loop................................................................................... 7-10
7.11
Checking Contact I/O....................................................................................... 7-10
7.11.1
Checking Contact Output.................................................................. 7-10
7.11.2
Checking Calibration Contact Output............................................... 7-11
7.11.3
Checking Input Contacts................................................................... 7-11
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
Toc-4
7.12 Calibration........................................................................................................ 7-12
8.
7.12.1
Calibration Setup.............................................................................. 7-12
7.12.2
Manual Calibration............................................................................ 7-13
Detailed Data Setting................................................................................ 8-1
8.1
8.2
8.3
8.4
8.5
8.6
Current Output Setting...................................................................................... 8-1
8.1.1
Oxygen Analyzer - Setting Minimum Current (4 mA) and Maximum
Current (20 mA).................................................................................. 8-1
8.1.2
Humidity Analyzer - Setting Minimum Current (4 mA) and Maximum
Current (20 mA).................................................................................. 8-1
8.1.3
Input Ranges....................................................................................... 8-1
8.1.4
Setting Output Smoothing Coefficient................................................ 8-4
8.1.5
Selection of Output Mode................................................................... 8-4
8.1.6
Default Values..................................................................................... 8-4
Output Hold Setting........................................................................................... 8-4
8.2.1
Definition of Equipment Status........................................................... 8-5
8.2.2
Preference Order of Output Hold Value.............................................. 8-5
8.2.3
Output Hold Setting............................................................................. 8-6
8.2.4
Default Values..................................................................................... 8-6
Alarms Setting.................................................................................................... 8-6
8.3.1
Setting the Alarm Values..................................................................... 8-7
8.3.2
Alarm Output Actions.......................................................................... 8-7
8.3.3
Alarm Setting Procedure..................................................................... 8-8
8.3.4
Default Values..................................................................................... 8-9
Output Contact Setup........................................................................................ 8-9
8.4.1
Output Contact.................................................................................... 8-9
8.4.2
Setting Procedure............................................................................. 8-10
8.4.3
Default Values................................................................................... 8-12
Input Contact Settings..................................................................................... 8-12
8.5.1
Input Contact Functions.................................................................... 8-12
8.5.2
Setting Procedure............................................................................. 8-13
8.5.3
Default Values................................................................................... 8-14
Other Settings.................................................................................................. 8-14
8.6.1
Setting the Date-and-Time................................................................ 8-14
8.6.2
Setting Periods over which Average Values Are Calculated and Periods
over which Maximum and Minimum Values Are Monitored............. 8-14
8.6.3
Setting Fuels..................................................................................... 8-15
8.6.4
Setting Measurement Gas Temperature and Pressure.................... 8-19
8.6.5
Setting Purging................................................................................. 8-21
8.6.6
Setting Passwords............................................................................ 8-22
9. Calibration.................................................................................................. 9-1
9.1
Calibration Briefs............................................................................................... 9-1
9.1.1
Principle of Measurement with a zirconia oxygen analyzer............... 9-1
9.1.2
Measurement Principle of Zirconia Humidity Analyzer....................... 9-2
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
Toc-5
9.1.3
Calibration Gas................................................................................... 9-4
9.1.4 Compensation..................................................................................... 9-4
9.1.5
9.2
Characteristic Data from a Sensor Measured During Calibration...... 9-5
Calibration Procedures..................................................................................... 9-6
9.2.1 Mode................................................................................................... 9-6
9.2.2
Calibration Procedure......................................................................... 9-7
9.2.3
Zero gas Concentration...................................................................... 9-7
9.2.4
Span gas Concentration..................................................................... 9-7
9.2.5
Setting Calibration Time...................................................................... 9-8
9.3 Calibration.......................................................................................................... 9-9
10.
9.3.1
Manual Calibration.............................................................................. 9-9
9.3.2
Semi-automatic Calibration.............................................................. 9-10
9.3.3
Automatic Calibration........................................................................ 9-10
Other Functions....................................................................................... 10-1
10.1
Detailed-data Display....................................................................................... 10-1
10.1.1
Span gas and Zero gas Correction Ratios....................................... 10-1
10.1.2
Cell Response Time.......................................................................... 10-2
10.1.3
Robustness of a Cell......................................................................... 10-2
10.1.4
Cell Temperature............................................................................... 10-2
10.1.5
C. J. Temperature............................................................................. 10-2
10.1.6
Cell Voltage....................................................................................... 10-3
10.1.7
Thermocouple Voltage...................................................................... 10-3
10.1.8
Cold Junction Resistance (C.J. Voltage).......................................... 10-3
10.1.9
Cell’s Internal Resistance................................................................. 10-3
10.1.10
Software Revision............................................................................. 10-3
10.1.11
Maximum Oxygen Concentration, Humidity and Mixing Ratio......... 10-4
10.1.12
Minimum Oxygen Concentration, Humidity and Mixing Ratio.......... 10-4
10.1.13
Average Oxygen Concentration, Humidity and Mixing Ratio........... 10-4
10.1.14
Heater On-Time Ratio....................................................................... 10-4
10.1.15 Time.................................................................................................. 10-4
10.2
10.3
10.1.16
History of Calibration Time................................................................ 10-4
10.1.17
Power Supply Voltage....................................................................... 10-4
10.1.18
Power Frequency.............................................................................. 10-4
Trend Graph...................................................................................................... 10-4
10.2.1
Setting Display Items........................................................................ 10-5
10.2.2
Sampling Interval.............................................................................. 10-5
10.2.3
Setting Upper and Lower Limit Values on Graph............................. 10-6
10.2.4
Default Setting.................................................................................. 10-6
Other Display-related Functions.................................................................... 10-7
10.3.1
Auto-Return Time.............................................................................. 10-7
10.3.2
Entering Tag Name........................................................................... 10-7
10.3.3
Language Selection.......................................................................... 10-7
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
Toc-6
10.4
Blow Back......................................................................................................... 10-8
10.4.1 Mode................................................................................................. 10-8
Operation of Blow back..................................................................... 10-8
10.4.3
Setting Output Hold Time and Blow back Time................................ 10-9
10.4.4
Setting Interval, Start Date, and Start Time...................................... 10-9
10.4.5
Default Setting................................................................................10-10
10.5
Parameter Initialization..................................................................................10-10
10.6
Reset................................................................................................................10-17
10.7
Handling of the ZO21S Standard Gas Unit..................................................10-17
10.8
11.
10.4.2
10.7.1
Standard Gas Unit Component Identification.................................10-18
10.7.2
Installing Gas Cylinders.................................................................. 10-19
10.7.3
Calibration Gas Flow.......................................................................10-19
Methods of Operating Valves in the ZA8F Flow Setting Unit....................10-21
10.8.1
Preparation Before Calibration....................................................... 10-22
10.8.2
Operating the Span Gas Flow Setting Valve.................................. 10-22
10.8.3
Operating the Zero Gas Flow Setting Valve...................................10-22
10.8.4
Operation After Calibration.............................................................10-23
Inspection and Maintenance.................................................................. 11-1
11.1
11.2
Inspection and Maintenance of the Detector................................................ 11-1
11.1.1
Cleaning the Calibration Gas Tube................................................... 11-1
11.1.2
Replacing the Sensor Assembly....................................................... 11-2
11.1.3
Replacement of the Heater Unit....................................................... 11-4
11.1.4
Replacement of Dust Filter............................................................... 11-6
11.1.5
Replacement of O-ring...................................................................... 11-6
11.1.6
Cleaning the High Temperature Probe Adapter................................ 11-7
11.1.7
Stopping and Re-starting Operation................................................. 11-7
Inspection and Maintenance of the Converter............................................. 11-8
11.2.1
Replacing Fuses............................................................................... 11-8
11.2.2 Cleaning............................................................................................ 11-9
11.2.3
11.3
12.
Adjust LCD screen contrast.............................................................. 11-9
Replacing Flowmeter in ZR40H Automatic Calibration Unit....................... 11-9
Troubleshooting...................................................................................... 12-1
12.1
12.2
12.3
Displays and Remedies When Errors Occur................................................ 12-1
12.1.1
Error Types........................................................................................ 12-1
12.1.2
Remedies When an Error Occurs..................................................... 12-2
Displays and Remedies When Alarms are Generated................................. 12-4
12.2.1
Alarm Types...................................................................................... 12-4
12.2.2
Remedies When Alarms are Generated........................................... 12-5
Countermeasures When Measured Value Shows Error............................ 12-11
12.3.1
Measured Value Higher (Lower for Humidity Analyzer)Than True Value.. 12-11
12.3.2
Measured Value Lower (Higher for Humidity Analyzer) Than True Value.12-12
12.3.3
Measurements Sometimes Show Abnormal Values......................12-13
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
Toc-7
Customer Maintenance Parts List.......................................CMPL 11M12A01-02E
Customer Maintenance Parts List.......................................CMPL 11M12C01-01E
Customer Maintenance Parts List.......................................CMPL 11M03B01-10E
Customer Maintenance Parts List.......................................CMPL 11M03B01-05E
Customer Maintenance Parts List.......................................CMPL 11M12A01-11E
Customer Maintenance Parts List...........................................CMPL 11M3D1-01E
Revision Information................................................................................................i
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
Blank Page
<1. Overview>
1-1
1. Overview
The EXAxt ZR Separate type Zircon Oxygen/Humidity Analyzer is used to monitor and control
the oxygen concentration in combustion gases, in boilers and industrial furnaces, for wide
application in industries which consume considerable energy — such as steel, electric power,
oil and petrochemical, ceramics, paper and pulp, food, or textiles, as well as incinerators and
medium/small boilers. It can help conserve energy in these industries.
The EXAxt ZR also contributes to preservation of the earth’s environment in preventing global
warming and air pollution by controlling complete combustion to reduce CO2, SOx and NOx.
The ZR22G Separate type Detector uses a high-reliability Zirconia sensor, and its heater
assembly can be replaced on site. The detector is mounted, for example, on the wall of a flue and
can measure the gases directly.
For use in combustion gases at temperatures up to 1400°C, choose the general use 0.15 m long
detector, which is combined with the ZO21P, the high temperature probe protector.
The EXAxt ZR Separate-type Zirconia Humidity Analyzer is used to measure the humidity of hot
gases continuously in driers which use hot gas or electricity as the heat source. It can also be
used in a variety of manufacturing applications in humidifiers, as well as in driers, for humidity
measurement and control. It can help improve productivity in these application fields.
The converter is equipped with an LCD touch screen which has various setting displays, a
calibration display, oxygen concentration trend display, with easier operation and improvement
of display functions. The converter is equipped with various standard functions such as
measurement and calculation as well as maintenance functions including self-test. Analyzer
calibration can also be fully automated — and the ZR40H, an automatic calibration unit,
is available. Choose the detector version which best suits your needs so that an optimal
combustion control system can be obtained.
Some examples of typical system configurations are illustrated below:
1.1
< EXAxt ZR > System Configuration
The system configuration should be determined by the conditions; e.g. whether calibration is
to be automated, and whether flammable gas is present and requires safety precautions. The
system configuration can be classified into three basic patterns as follows:
1.1.1
System 1
This is the simplest system consisting of a detector and a converter. This system can be
implemented for monitoring oxygen concentration in the combustion gases of a package
boiler, and can be implemented for monitoring humidity in a production process such as food
production.
No piping is required for the reference gas (air) which is fed in at the installation site. The handy
the ZO21S standard gas unit is used for calibration.
Zero gas from this unit and span gas (air) is sent to the detector through a tube which is
connected during calibration.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
1-2
<1. Overview>
ZR22G Separate type
Zirconia Oxygen/Humidity Analyzer, Detector
Stop valve
ZR402G Converter
EXA ZR402G
Signal
(6-core shield cable)
Heater(2-core)
100 to
~ 240 V AC
Contact input
Analog output, contact output
(Digital output HART)
Calibration gas
~
ZO21S Standard gas unit
F1-1E.ai
Figure1.1
NOTE
• As this system uses ambient air for the reference gas, measuring accuracy will be affected
by the installation location.
• A needle (stop) valve should be connected to the calibration gas inlet of the detector.
The valve should be fully closed unless calibration is in progress.
1.1.2
System 2
This system is for monitoring and controlling oxygen concentration in the combustion gases of a
large-size boiler or heating furnace. Clean (dry) air (21 vol%O2) is used as the reference gas and
the span gas for calibration.
In case of humidity analyzer, this system is for accurate monitoring and controlling humidity when
the installation environment is polluted with gases other than the air. Instrument air (clean and dry
air of oxygen concentration 21%) is used for the reference gas and the span gas for calibration.
Zero gas is fed in from a cylinder during calibration. The gas flow is controlled by the ZA8F flow
setting unit (for manual valve operation).
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
1-3
<1. Overview>
ZR402G Converter
ZR22G Separate type
Zirconia Oxygen/Humidity Analyzer, Detector
EXA ZR402G
Signal
(6-core shield cable)
Check valve
or Stop Valve
100 to
~ 240 V AC
Heater(2-core)
Flowmeter
Reference
gas
Needle
valve
Contact input
Analog output, contact output
(Digital output HART)
Air Set
Calibration gas
Instrument air
Span gas
(Same as Zero gas calibration unit)
Pressure
reducing
valve
ZA8F Flow Setting Unit
Zero gas
cylinder
Calibration gas
unit case
F1-2E.ai
Figure1.2
1.1.3
System 3
This example, System 3, represents typical applications in large boilers and heating furnaces,
where there is a need to monitor and control oxygen concentration. The reference gas and
calibration-time span gas are (clean, dry) instrument air. Zero gas is supplied from a gas cylinder.
System 3 uses the ZR40H automatic calibration unit, with auto-switching of the calibration gas.
A “combustible gas detected” contact input turns off power to the heater. There’s also contact
output from the converter that can be used to operate a purge gas valve to supply air to the
sensor.
ZR22G Separate type
Zirconia Oxygen/Humidity Analyzer, Detector
*2
Check valve
ZR402G Converter
Signal*1
(6-core shield cable)
Heater (2-core)
Reference gas
Calibration gas
Needle
Flowmeter valve
EXA ZR402G
~
100 to 240 V AC
Contact input
Analog output, contact output
Digital output (HART)
Air Set
lnstrument air
Zero gas cylinder
Pressure
ZR40H
reducing
Automatic Calibration Unit valve
*3
Calibration gas
F1-3E.ai
unit case
*1
*2
*3
Shield cable ; Use shielded signal cables, and connect the shield to the FG terminal of the converter.
Select the desired probe from the Probe Configuration table on page 1-4.
When a zirconia oxygen analyzer is used, 100% N2 gas cannot be used as the zero gas. Use approximately 1% of O2 gas (N2based).
Figure1.3
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
1-4
<1. Overview>
1.2
< EXAxt ZR > System Components
1.2.1
System Components
Separate type
Oxygen Humidity
System config.
Analyzer Analyzer
Ex.1 Ex.2 Ex.3
A
A
A
A
A
A
A
A
A
A
System Components
Model ZR22G Separate type Zirconia Oxygen Analyzers ,Detector
Model ZR402G Separate type Zirconia Oxygen Analyzer, Converter
Model ZO21P Adapter for High Temperature Probe of separate type
Zirconia Oxygen Analyzer
E7046EC, E7046EN Ejector Assembly for High temperature of separate
type Oxygen Analyzer
Model ZO21R Probe Protector for Zirconia Oxygen Analyzers
K9471UA Dust Filter for Oxygen Analyzer
K9471UC Dust Guard Protector
ZH21B Dust protector (only for Humidity Analyzer)
Model ZO21S Standard Gas Unit
Model ZA8F Flow Setting Unit for manual calibration
Model ZR40H Automatic Calibration Unit for Separate type Oxygen
Analyzer
L9852CB, G7016XH Stop Valve for Calibration gas line
K9292DN,K9292DS Check Valve for Calibration gas line
G7003XF/K9473XK, G7004XF/K9473XG Air Set
G7001ZC Zero gas Cylinder
G7013XF, G7014XF Pressure Reducing Valve for Gas Cylinder
E7044KF Case Assembly for Calibration gas Cylinder
Model ZR22A Heater Assembly (Spare Parts for Model ZR22G)
B
B
B
B
B
B
B
B
B
B
B
B
A
B
B
B
B
B
B
B
B
B
B
B
B
B
B
B
B
B
A
B
B
A
A
A
A
A
B
B
B
B
B
B
B
B
B
B
B
B
B
B
B
A
A
B
(A)
(A)
A
A
A
A
B
B
A: Items required for the above system example
B: To be selected depending on each application. For details, refer to corresponding chapter.
(A): Select either
1.2.2
Detectors and Accessories
Sample gas temperature 0 to 700°C
Mounting
Horizontal
to
vertical
Vertical
Insertion
length
Detector
(ZR22G)
0.15
to
2m
2.5 m
or more
3m
or less
Horizontal
to
vertical
0.15
to
2m
2.5 m
or more
High temperature detector
Application
Boiler
Heating
furnace
Detector
(only ZR22G
-040)
Dust protector
(ZH21B)
Sample
outlet
Probe Protector
(ZO21R)
Detector
(ZR22G)
Sample inlet
Detector(ZR22G)
or
Dust guard
protector
(K9471UC)
High temperature
detector
Sample inlet
Temperature:
Probe material: SUS310S 800˚C
Probe material: SiC 1400˚C
Mounting: Vertical downwards
Insertion length: 1.0 m, 1.5 m
When duct pressure is atmospheric
or negative, attach air ejector.
High temperature auxiliary ejector
assembly (E7046EC, E7046EN)
For pulverized
coal boiler
with gas flow
velocity
10 m/sec or
more
Pressure gauge
Black liquid
recovery boiler
Dust filter for
Oxygen Analyzer
(K9471UA)
Application
Heating
furnace
Absorption
structure
Probe adapter
for high
temperature use
ZO21P-H
Humidity analyzer use
Gas Flow
Horizontal
to
vertical
Vertical
General use Probe
Sample gas temperature 700 to 1400˚C
Cement
Kiln
Needle
valve
Ejector
Inlet
+
Blow
F06.ai
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
2.
<2. Specifications>
2-1
Specifications
This chapter describes the specifications for the following:
ZR22G
General use separate type detector (See Section 2.2.1)
ZO21R
Probe protector (See Section 2.2.2)
ZH21B
Dust protector (See Section 2.2.3)
ZR22G (0.15 m)
High temperature separate type detector (See Section 2.3.1)
ZO21P
Adapter for High temperature probe (See Section 2.3.2)
ZR402G
Separate type converter (See Section 2.4)
ZA8F
Flow setting unit (See Section 2.5.1)
ZR40H
Automatic calibration unit (See Section 2.5.2)
ZO21S
Standard gas unit (See Section 2.6)
2.1
General Specifications
Standard Specifications
Measured Object: Oxygen concentration in combustion exhaust gas and mixed gas
(excluding inflammable gases, may not be applicable corrosive gas such
as ammonia, chlorine is present-check with YOKOGAWA)
The sampling gases containing a corrosive gas such as ammonia or
chlorine may be applicable to our oxygen gas analyzer. In this case,
contact with YOKOGAWA and its agency.
(In case of Humidity Analyzer, Water vapor (in vol%) in mixed gases (air
and water vapor))
Measurement System: Zirconia system
Measurement Range: 0.01 to 100 vol%O2
(In case of Humidity Analyzer, 0 to 100 vol% H2O or 0 to 1.000 kg/kg)
Output Signal:
4 to 20 mA DC (maximum load resistance 550 Ω)
Setting Range: Any setting in the range of 0 to 5 through 0 to 100 vol%O2 (in 1 vol%O2),
or partial range
In case of Humidity Analyzer,
Moisture quantity; 0 to 25 through 0 to 100 vol% H2O (in 1 vol%
H2O), or partial range.
Mixture ratio; 0 to 0.2 through 0 to 1.000 kg/kg (in 0.001 kg/kg), or
partial range.
Digital Communication (HART): 250 to 550 Ω, depending on number of field devices connected
to the loop (multi-drop mode).
Note: HART is a registered trademark of the HART Communication Foundation.
Display Range: Oxygen concentration; 0 to 100 vol%O2
In case of Humidity Analyzer,
Moisture quantity; 0 to 100 vol% H2O, Mixture ratio; 0 to 1 kg/kg, Relative
humidity; 0 to 100% RH (Note), Dew point; -40 to 370°C (Note)
Note: These values are affected by temperature and absolute pressure, So accurate temperature and
pressure values must be inputted to the converter.
Warm-up Time:
Approx. 20 min.
These characteristics are calculated by oxygen concentration measured in air which include
water vapor.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<2. Specifications>
2-2
Repeatability: (Excluding the case where the reference gas is by natural convection)
±0.5% Maximum value of set range. (range from 0 to 5 vol%O2 or more
and less than 0 to 25 vol%O2 range)
±1% Maximum value of set range. (range from 0 to 25 vol%O2 or more
and up to 0 to 100 vol%O2 range)
In case of Humidity Analyzer, ± 1 vol% H2O; (Sample gas pressure 2 kPa
or less)
Linearity: (Excluding standard gas tolerance)
(Excluding the case where the reference gas is by natural convection)
(Use oxygen of known concentration (within the measuring range) as the
zero and span calibration gases.)
±1% Maximum value of set range ; 0 to 5 vol%O2 or more and less than
0 to 25 vol%O2 range (Sample gas pressure: within ±4.9 kPa)
±3% Maximum value of set range ; 0 to 25 vol%O2 or more and less than
0 to 50 vol%O2 range (Sample gas pressure: within ±0.49 kPa)
±5% Maximum value of set range ; 0 to 50 vol%O2 or more and up to 0 to
100 vol%O2 range (Sample gas pressure: within ±0.49 kPa)
In case of Humidity Analyzer
± 2 vol% H2O; (Sample gas pressure: within ± 0.49 kPa)
± 3 vol% H2O; (Sample gas pressure: 2 kPa or less)
Drift: (Excluding the first two weeks in use)
(Excluding the case where the reference gas is by natural convection.)
Both zero and span ±2% Maximum value of set range/month
In case of Humidity Analyzer
Both zero and span ± 3 vol% H2O/month
Response Time :
Response of 90% within 5 seconds. (Measured after gas is introduced
from calibration gas inlet and analog output start changing.)
Safety and EMC conforming standards the ZR22G and ZR402G
Installation altitude based on IEC 61010:
2000 m or less
Category based on IEC 61010: II (Note)
Pollution degree based on IEC 61010: 2 (Note)
Note: Installation category, called over-voltage category, specifies impulse withstand voltage.
Category II is for electrical equipment.
Pollution degree indicates the degree of existence of solid, liquid, gas or other inclusions
which may reduce dielectric strength. Degree 2 is the normal indoor environment.
Safety:
EMC:
EN 61010-1, EN 61010-2-030, CAN/CSA-C22.2 No. 61010-1,
UL Std. No. 61010-1
EN 61326-1 Class A*, Table 2,
EN 61326-2-3, EN 61000-3-2
*: Influence of immunity environment (Criteria A ): ±20% of F. S.
EMC Regulatory Arrangement in Australia and New Zealand (RCM)
EN61326-1 Class A
Korea Electromagnetic Conformity Standard
RoHS:
EN 50581
CAUTION
This instrument is a Class A product, and it is designed for use in the industrial environment.
Please use this instrument in the industrial environment only.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
2.2
<2. Specifications>
2-3
General use Separate type Detector and
Related Equipment
General use separate type detector ZR22G can be used in combination with the probe protector
ZO21R (see Section 2.2.2).
In case of Humidity Analyzer, the “Detector with dust protector” consists of ZR22G general-use
separate-type detector and ZH21B dust protector (refer to Section 2.2.3).
2.2.1
ZR22G General use Separate type Detector
Sample Gas Temperature:0 to 700°C (Probe only)
It is necessary to mount the cell using Inconel cell-bolts when the
temperature is greater than 600°C.
700 to 1400°C (with High Temperature Probe Adapter)
For high temperature sample gas (700 to 1400°C), apply 0.15 m long
probe and High Temperature Probe Adapter ZO21P.
Sample Gas Pressure: -5 to 250 kPa (When the pressure in the furnace exceeds 3 kPa, it is
recommended to use pressure compensated type. When the pressure
in the furnace exceeds 5 kPa, pressure compensated type is required.)
For 0.15 m probe, -0.5 to 5 kPa. No pressure fluctuation in the furnace
should be allowed.
No pressure fluctuation in the process should be allowed.
In case of Humidity Analyzer, -5 to 20 kPa
Note: When the detector is used in conjunction with a check valve and a ZA8F Flow Setting Unit, the
maximum pressure of sample gas is 150 kPa. When with a check valve and a ZR40H Automatic
Calibration Unit, it is 200 kPa. If the pressure of your sample gas exceeds these limits, consult with
Yokogawa.
Probe Length: 0.15, 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0, 3.6, 4.2, 4.8, 5.4 m
In case of Humidity Analyzer, 0.15 m is exclided
Probe Material:
SUS 316 (JIS)
Ambient Temperature: -20 to 150°C
Reference Gas System: Natural Convection, Instrument Air, or Pressure Compensation(other
than for probe length 0.15 m)
Instrument Air System (excluding Natural Convection) :
Pressure; 200 kPa + the pressure inside the furnace (It is recommended
to use air which has been dehumidified by cooling to dew point -20°C or
less, and dust or oil mist removed.)
Consumption; Approx. 1 Nl/min
Wetted Material: SUS 316 (JIS), Zirconia, SUS 304 (JIS) (flange), Hastelloy B, (Inconel
600, 601)
Construction: Heater and thermocouple replaceable construction.
Non explosion-proof JIS C0920/equivalent to IP44D.
Equivalent to NEMA 4X/IP66 (Achieved when the cable entry is
completely sealed with a cable gland in the recirculation pressure
compensated version.)
Terminal Box Case: Material; Aluminum alloy
Terminal Box Paint Color: Case; Mint green (Munsell 5.6BG3.3/2.9)
Cover; Mint green (Munsell 5.6BG3.3/2.9)
Finish:
Polyurethane corrosion-resistant coating
Gas Connection:
Rc1/4 or 1/4FNPT
Wiring Connection:
G1/2, Pg 13.5, M20 × 1.5, 1/2 NPT
Installation:
Flange mounting
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<2. Specifications>
2-4
Probe Mounting Angle: Horizontal to vertically downward.
When the probe insertion length is 2 m or less, installing at angles from
horizontal to vertically downward is available.
When the probe insertion length is exceeds 2.5 m, mount vertically
downward (within ±5°), and use a probe protector.
Weight: Insertion length of 0.4 m: approx. 6 kg (JIS 5K 65) / approx. 11 kg (ANSI 150 4)
Insertion length of 1.0 m: approx. 8 kg (JIS 5K 65) / approx. 13 kg (ANSI 150 4)
Insertion length of 1.5 m: approx. 10 kg (JIS 5K 65) / approx. 15 kg (ANSI 150 4)
Insertion length of 2.0 m: approx. 12 kg (JIS 5K 65) / approx. 17 kg (ANSI 150 4)
Insertion length of 3.0 m: approx. 15 kg (JIS 5K 65) / approx. 20 kg (ANSI 150 4)
Insertion length of 3.6 m: approx. 17 kg (JIS 5K 65) / approx. 22 kg (ANSI 150 4)
Insertion length of 4.2 m: approx. 19 kg (JIS 5K 65) / approx. 24 kg (ANSI 150 4)
Insertion length of 4.8 m: approx. 21 kg (JIS 5K 65) / approx. 26 kg (ANSI 150 4)
Insertion length of 5.4 m: approx. 23 kg (JIS 5K 65) / approx. 28 kg (ANSI 150 4)
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
2-5
<2. Specifications>
l Model and Codes
Style : S2
Model
Suffix code
Option code
Description
ZR22G
----------------------------------
----------
Separate type Zirconia Oxygen/Humidity Analyzer, Detector
Length
-015
-040
-070
-100
-150
-200
-250
-300
-360
-420
-480
-540
-------------------------------------------------------------------------------------------------------------
0.15 m (for high temperature use)
0.4 m
0.7 m
1.0 m
1.5 m
2.0 m
2.5 m
(*2)
3.0 m
(*2)
3.6 m
(*2)
4.2 m
(*2)
4.8 m
(*2)
5.4 m
(*2)
-------------------
Stainless steel
Stainless steel with Inconel calibration gas tube
-------------------------------------------------------------------------------------------------------------------------------
ANSI Class 150 2 RF SUS304 (JIS)
ANSI Class 150 3 RF SUS304 (JIS)
ANSI Class 150 4 RF SUS304 (JIS)
DIN PN10 DN50 A SUS304 (JIS)
DIN PN10 DN80 A SUS304 (JIS)
DIN PN10 DN100 A SUS304 (JIS)
JIS 5K 65 FF SUS304 (JIS)
JIS 10K 65 FF SUS304 (JIS)
JIS 10K 80 FF SUS304 (JIS)
JIS 10K 100 FF SUS304 (JIS)
JIS 5K 32 FF SUS304 (JIS) (for high temperature use) (*4)
JPI Class 150 4 RF SUS304 (JIS)
JPI Class 150 3 RF SUS304 (JIS)
Westinghouse
----------------------------
Natural convection
External connection (Instrument air)
Pressure compensated
(*11)
-------------------
Rc 1/4
1/4 NPT(Female)
----------------------------------------------
G1/2
Pg13.5
M20 x1.5 mm
1/2NPT
Quick connect
----------------------------
Japanese
English
Chinese
----------
Always -A
Wetted material
-S
-C
Flange
(*3)
-A
-B
-C
-E
-F
-G
-K
-L
-M
-P
-Q
-R
-S
-W
Reference gas
-C
-E
-P
Gas Thread
-R
-T
Connection box thread
-P
-G
-M
-T
-Q
Instruction manual
-J
-E
-C
—
-A
Options
*1
*2
*3
*4
*5
*6
*7
*8
*9
*10
*11
(*10)
(*11)
(*9)
/C
Inconel bolt
(*5)
Valves
/CV
/SV
Check valve
Stop valve
(*6)
(*6)
Filter
/F1
/F2
Dust Filter
(*7)
Dust Guard Protector (*7)
/SCT
/PT
Stainless steel tag plate
Printed tag plate
Tag plates
(*1)
(*8)
(*8)
Used with the ZO21P High Temperature Probe Adapter. Select flange (-Q). Only for Oxygen Analyzer.
When installing horizontally the probe whose insertion length is 2.5 meters or 3.0 meters, use the Probe Protector. Be sure to
specify ZO21R-L-200- . Specify the flange suffix code either -C or -K.
The thickness of the flange depends on its dimensions.
Not used in conjunction with -P (pressure compensation) for reference gas. The flange thickness does not conform to JIS
specification
Inconel probe bolts and U shape pipe are used. Use this option for high temperature use (ranging from 600 to 700 °C).
Specify either /CV or /SV option code.
Not used with the high temperature humidity analyzer.
Specify either /SCT or /PT option code.
Not waterproof, avoid rain. Operating maximum temperature is 80°C. Available only in the U.S.
Recommended if sample gas contains corrosive gas like chlorine.
Piping for reference gas must be installed to supply reference gas constantly at a specified flow rate.
□
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
2-6
<2. Specifications>
l EXTERNAL DIMENSIONS
1. Model ZR22G Separate type Zirconia Oxygen/Humidity Analyzer, Detectors
L
283 to 292
Unit : mm
Ø124±3
t
85
Ø50.8
L=0.15, 0.4, 0.7, 1.0,
1.5, 2.0, 2.5, 3.0
3.6, 4.2, 4.8, 5.4 (m)
Rc1/4 or 1/4NPT
Reference gas inlet
155 to 163
69
2-G1/2,2-1/2NPT etc.
Cable connection port
C
ØA
48
25
ØB
Rc1/4 or 1/4NPT
Calibration gas inlet
Flange
Flange
ANSI Class 150 2 RF SUS304
ANSI Class 150 3 RF SUS304
ANSI Class 150 4 RF SUS304
DIN PN10 DN50 A SUS304
DIN PN10 DN80 A SUS304
DIN PN10 DN100 A SUS304
JIS 5K 65 FF SUS304
JIS 10K 65 FF SUS304
JIS 10K 80 FF SUS304
JIS 10K 100 FF SUS304
JIS 5K 32 FF SUS304
JPI Class 150 4 RF SUS304
JPI Class 150 3 RF SUS304
Westinghouse
A
152.4
190.5
228.6
165
200
220
155
175
185
210
115
229
190
155
B
120.6
152.4
190.5
125
160
180
130
140
150
175
90
190.5
152.4
127
C
4 - Ø19
4 - Ø19
8 - Ø19
4 - Ø18
8 - Ø18
8 - Ø18
4 - Ø15
4 - Ø19
8 - Ø19
8 - Ø19
4 - Ø15
8 - Ø19
4 - Ø19
4 - Ø11.5
t
19
24
24
18
20
20
14
18
18
18
5
24
24
14
Flange
C
ØA
ØB
Flange
F2-1E.ai
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
2-7
<2. Specifications>
2. Model ZR22G...-P (with pressure compensation) Separate type Zirconia Oxygen
Analyzer, Detectors
303
t
85
Ø50.8
ø124±3
L
L=0.15, 0.4, 0.7, 1.0,
1.5, 2.0, 2.5, 3.0
3.6, 4.2, 4.8, 5.4 (m)
156
Rc1/4 or 1/4NPT
Reference gas inlet
87
2-G1/2, 2-1/2NPT etc.
Cable connection port
C
48
25
ØA
ØB
Reference gas outlet
Flange
PIPING
:B
PIPING : A
Flange
ANSI Class 150 2 RF SUS304
ANSI Class 150 3 RF SUS304
ANSI Class 150 4 RF SUS304
DIN PN10 DN50 A SUS304
DIN PN10 DN80 A SUS304
DIN PN10 DN100 A SUS304
JIS 5K 65 FF SUS304
JIS 10K 65 FF SUS304
JIS 10K 80 FF SUS304
JIS 10K 100 FF SUS304
JPI Class 150 4 RF SUS304
JPI Class 150 3 RF SUS304
Westinghouse
2.2.2
A
152.4
190.5
228.6
165
200
220
155
175
185
210
229
190
155
B
120.6
152.4
190.5
125
160
180
130
140
150
175
190.5
152.4
127
Rc1/4 or 1/4NPT
Calibration gas inlet
Stop Valve
Flange
C
C
4 - Ø19
4 - Ø19
8 - Ø19
4 - Ø18
8 - Ø18
8 - Ø18
4 - Ø15
4 - Ø19
8 - Ø19
8 - Ø19
8 - Ø19
4 - Ø19
4 - Ø11.5
t
19
24
24
18
20
20
14
18
18
18
24
24
14
PIPING
A
B
B
A
B
B
A
A
B
B
B
B
A
ØA
ØB
Flange
F2-2E.ai
ZO21R Probe Protector
Used when sample gas flow velocity is approx. 10 m/sec or more and dust particles wears the
detector in cases such as pulverized coal boiler of fluidized bed furnace (or burner) to protect
the detector from wearing by dust particles. When probe insertion length is 2.5 m or more and
horizontal installation, specify the ZO21R-L-200- *B to reinforce the probe.
Insertion Length:
1.05 m, 1.55 m, 2.05 m.
Flange: JIS 5K 65A FF equivalent. ANSI Class 150 4 FF (without serration)
equivalent. However, flange thickness is different.
Material:
SUS316 (JIS), SUS304 (JIS) (Flange)
Weight:
1.05 m; Approx. 6/10/8.5 kg (JIS/ANSI),
1.55 m; Approx. 9/13/11.5 kg (JIS/ANSI),
2.05 m; Approx. 12/16/14.5 kg (JIS/ANSI)
Installation: Bolts, nuts, and washers are provided for detector, probe adapter and
process-side flange.
□
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<2. Specifications>
2-8
l Model and Codes
Model
Suffix code
ZO21R
-L
Insertion
length
Flange ( *1)
Description
------
Probe Protector(0 to 700 °C)
-100
-150
-200
----------------
1.05 m (3.5 ft)
1.55 m (5.1 ft)
2.05 m (6.8 ft)
-J
-A
-----------
JIS 5K 65 FF SUS304
ANSI Class 150 4 FF SUS304
------
Style B
Style code
*1
Option
code
*B
Thickness of flange depends on dimensions of flange.
Flange <1>
(with bolts, nuts and washer)
gasket (t3.0)
ØA
Ø60.5
Gas flow
Washer (12)
Mounting nut (M12)
SUS316
ØB
D
t
l (Insert length)
ØB
C
l=1050,1550,2050
Dimensions of
holes on opposing surface
F2-3E.ai
Flange<1>
A
B
C
t
D
155
130
4 - Ø15
5
40
228.6
190.5
8 - Ø19
12
50
JIS 5K 65 FF SUS304
ANSI Class 150 4 FF SUS304
2.2.3
ZH21B Dust Protector
This protector is designed to protect the probe output from dust agitation (i.e., to prevent
combustible materials from entering the probe cell where humidity measurements are made) in a
dusty environment.
Insertion length:
0.428m
Flange: JIS 5K 80 FF SUS304 or ANSI Class 150 4 FF SUS304 (However,
flange thickness is different)
Material:
SUS 316(JIS), SUS 304(JIS) (flange)
Weight:
Approximately 6kg (JIS), approximately 8.5kg (ANSI)
Mounting: Mounted on the probe or process flange with bolts and associated nuts
and washers.
l Model and Codes
Model
ZH21B
Suffix code
Option
code
Description
-------------
------
Dust Protector (0 to 600 °C)
------
0.428 m
-----------
JIS 5K 80 FF SUS304 (*1)
ANSI Class 150 4B FF SUS304 (*2)
------
Style B
Insertion length
Flange ( *1)
Style code
-40
-J
-A
*B
Note: The flange thickness varies.
*1:
Specify the probe ZR22G-040-□-K
*2:
Specify the probe ZR22G-040-□-C
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<2. Specifications>
2-9
Unit: mm
428 (Insertion length)
øA
øB
ø72
ø76.3
t
JIS flange
Install facing upwards
D
ANSI flange
øB
øB
C
C
ZH21B.ai
Flange
JIS 5K 80 FF SUS304
ANSI Class 150 4B FF SUS304
2.3
A
B
C
t
D
180
145
4 - Ø19
12
40
228.6
190.5
8 - Ø19
12
50
Separate type Detector for High
Temperature and Related Equipment
2.3.1
ZR22G (0.15m) Separate type Detector for High
Temperature
Standard Specifications
Construction:
Water-resistant, non-explosionproof
Probe length:
0.15 m
Terminal box:
Aluminum alloy
Probe material:
Probe material in contact with gas: SUS 316 (JIS) (Probe), SUS 304
(JIS) (Flange), Zirconia (Sensor), Hastelloy B, (Inconel 600, 601)
Weight:
Approx. 3 kg
Installation:
Flange mounting (The use of high temperature detector probe adapter
ZO21P is necessary.)
Flange standard:
JIS 5 K 32 FF equivalent (thickness varies)
Mounting angle:
Any angle between horizontal and vertical (high temperature probe is
fitted with an adapter)
Reference gas and calibration gas piping connection: Rc 1/4 or 1/4 NPT female
Cable inlet:
G 1/2, Pg 13.5, M20 x 15, 1/2 NPT
Ambient temperature: -20 to 150°C
Sample gas temperature: 0 to 700°C (temperature at the measuring point of the sampling gas.
0 to 750°C or 0 to 1400°C when the probe adapter for high
temperature is used.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<2. Specifications>
2-10
Temperature of the probe adapter shall not exceed 300°C to protect the
gasket and avoid the bolts seizing together.
Sample gas pressure: -0.5 to 5 kPa: when used at the range of more than 0 to 25 vol%O2,
-0.5 to 0.5 kPa. (An ejector assembly is required for negative pressure
application.)
Model and Code:
Refer to “Model and Codes” in page 2-5.
External Dimensions: Refer to the Figure in page 2-6.
2.3.2
ZO21P High Temperature Probe Adapter
Measuring O2 in the high temperature gases (exceeds 700°C) requires a general use probe
ZR22G of 0.15 m length and a high temperature probe adapter.
Sample gas temperature: 0 to 1400°C (when using SiC probe)
0 to 800°C (when using SUS 310S probe adapter)
Sample gas pressure: -0.5 to 5 kPa. When using in the range of 0 to 25 vol%O2 or more,
the sample gas pressure should be in the range of -0.5 to 0.5 kPa.
(Where the sample gas pressure for the high temperature probe is
negative, an ejector assembly is necessary.)
Insertion length:
0.5, 0.6, 0.7, 0.8, 0.9, 1.0, 1.5 m
Material in Contact with Gas: SUS 316 (JIS), SiC or SUS 310S, SUS 304 (JIS) (flange)
Probe Material:
SiC, SUS 310S (JIS)
Installation:
Flange mounting (FF type or RF type)
Probe Mounting Angle: Vertically downward within ± 5° Where the probe material is SUS 310S,
horizontal mounting is available.
Construction:
Non explosion-proof. Rainproof construction
Weight(example):
Insertion length of 1.0 m: approx. 5.3 kg (JIS) / approx. 11.3 kg (ANSI)
Insertion length of 1.5 m: approx. 5.8 kg (JIS) / approx. 11.8 kg (ANSI)
l Model and Codes
Model
ZO21P
Material
Insertion
length
Flange
Style code
Option
Suffix code
Option code
-H
-A
-B
-050
-060
-070
-080
-090
-100
-150
-J
-N
-M
-L
-A
-R
-Q
-T
-S
-E
*B
Ejector
Tag plate
Description
-------
High Temperature Probe Adapter
-------------
SiC
SUS 310S (JIS)
------------------------------------
0.5 m
0.6 m
0.7 m
0.8 m
0.9 m
1.0 m
1.5 m
-------------------------------------------------------------
JIS 5K 50 FF SUS304 (JIS)
JIS 10K 65 FF SUS304 (JIS)
JIS 10K 80 FF SUS304 (JIS)
JIS 10K 100 FF SUS304 (JIS)
ANSI Class 150 4 RF SUS304 (JIS)
ANSI Class 150 2 1/2 RF SUS304 (JIS)
ANSI Class 150 3 RF SUS304 (JIS)
JPI Class 150 3 RF SUS304 (JIS)
JPI Class 150 4 RF SUS304 (JIS)
DIN PN10 DN50 A SUS304 (JIS)
-------
Style B
/EJ1
/EJ2
/SCT
Ejector Assy with E7046EC
Ejector Assy with E7046EN
Stainless steel tag plate
Note: For this high-temperature use probe adapter, be sure to specify the ZR22G probe of its insertion length 0.15 meters.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<2. Specifications>
Unit: mm
Approx. 351
Sample gas outlet
2-11
Flange (Thickness 5)
JIS 5K 32 FF equivalent
180
Rc1/2(Note2)
Ø124±3
110
ØA
85
Ø115
Ø60.5
170
Ø60.5
t
Approx. 100
Flange <1>
Detector(ZR22G)
Flange provided
by customer
Approx. 48
Ø52 over
Reference gas inlet Rc1/4 or 1/4NPT
69
High temperature
Probe SiC pipe
Ø30
48
25
L (Insertion length) (Note1)
Approx. 215
Gasket (Thickness 1.5)
Pipe hole (2- G1/2, 2-1/2NPT, etc)
Calibration gas inlet Rc1/4 or 1/4NPT
C
ØA
(Note 1) L= 0.5、0.6、0.7、0.8、0.9、
1.0、1.5(m)
(Note 2) Sample gas outlet
(if the sample gas pressure is negative,
connect the auxiliary ejector.)
ØB
F2-4E.ai
Flange<1>
2.4
A
B
C
t
JIS 5K 50 FF SUS304
130
105
4 - Ø15
14
JIS 10K 65 FF SUS304
175
140
4 - Ø19
18
JIS 10K 80 FF SUS304
185
150
8 - Ø19
18
JIS 10K 100 FF SUS304
210
175
8 - Ø19
18
ANSI Class 150 4 RF SUS304
228.6
190.5
8 - Ø19
24
ANSI Class 150 3 RF SUS304
190.5
152.4
4 - Ø19
24
ANSI Class 150 2 RF SUS304
177.8
120.6
4 - Ø19
19
JPI Class 150 4 RF SUS304
229
190.5
8 - Ø19
24
JPI Class 150 3 RF SUS304
190
152.4
4 - Ø19
24
DIN PN10 DN50A SUS304
165
125
4 - Ø18
18
ZR402G Separate type Converter
l Standard Specification
The ZR402G Separate type Converter can be controlled by LCD touchscreen on the converter.
Display:
LCD display of size 320 by 240 dot with touchscreen.
Output Signal:
4 to 20 mA DC, two points (maximum load resistance 550 Ω)
Contact Output Signal: Four points (one is fail-safe, normally open)
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<2. Specifications>
2-12
Contact Input:
Two points
Analog Input:
One point (thermal input 4-20 mA)
Automatic Calibration Output: Two points (for dedicated automatic calibration unit)
Ambient Temperature: -20 to 55°C
Storage Temperature: -30 to 70°C
Ambient Humidity:
0 to 95%RH (Non-condensing)
Power Supply Voltage: Ratings; 100 to 240 V AC, Acceptable range; 85 to 264 V AC
Power Supply Frequency: Ratings; 50/60 Hz, Acceptable range; 45 to 66 Hz
Power Consumption: Max. 300 W, approx. 100 W for ordinary use.
Maximum Distance between Detector and Converter: Conductor two-way resistance must be
10 Ω or less (when a 1.25 mm2 cable or equivalent is used, 300 m or
less.)
Construction:
Outdoor installation, equivalent to NEMA 4X/IP66 (with conduit holes
completely sealed with a cable gland)
Wiring Connection:
G1/2, Pg13.5, M20 x 1.5 mm, 1/2 NPT (with plug), eight holes
Installation:
Panel, wall or pipe mounting
Case:
Aluminum alloy
Paint Color:
Door: Sliver gray (Munsell 3.2PB7.4/1.2)
Case: Sliver gray (Munsell 3.2PB7.4/1.2)
Finish:
Polyurethane corrosion-resistance coating
Weight:
Approx. 6 kg
l Functions
Display Functions: (inclused Humidity Analyzer)
Value Display; Displays values of the measured oxygen concentration, moisture
quantity, mixture ratio, etc
Graph Display; Displays trends of measured oxygen concentration, moisture quantity,
mixture ratio, etc.
Data Display; Displays various useful data for maintenance, such as cell temperature,
reference junction temperature, maximum/minimum oxygen
concentration, maximum/minimum moisture quantity or the like
Status Message; Indicates an alarm or error occurrence by flashing of the corresponding
icon. Indicates status such as warming-up, calibrating,or the like by
icons.
Alarm, Error Display; Displays alarms such as “Abnormal oxygen concentration”,
“Abnormal moisture quantity” or errors such as “Abnormal cell
e.m.f.” when any such status occurs.
Calibration Functions:
Automatic calibration; Requires the ZR40H Automatic Calibration Unit. It calibrates
automatically at specified intervals.
Semi-auto Calibration; Requires the ZR40H Automatic Calibration Unit. Input calibration
direction on the touchscreen or contact, then it calibrates
automatically afterwards.
Manual Calibration; Calibration with opening/closing the valve of calibration gas in
operation interactively with an LCD touchscreen.
Blow back Function:
Output through the contact in the set period and time. Auto/Semi_Auto selectable.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<2. Specifications>
2-13
Maintenance Functions:
Can operate updated data settings in daily operation and checking. Display data settings,
calibration data settings, blow back data settings, current output loop check, input/output
contact check.
Setup Functions:
Initial settings suit for the plant conditions when installing the converter.
Equipment settings, current output data settings, alarm data settings, contact data
settings, other settings.
Self-diagnosis:
This function diagnoses conditions of the converter or the detector and indicates when
any abnormal condition occurs.
Password Functions:
Enter your password to operate the analyzer excepting data display.
Individual passwords can be set for maintenance and setup.
Display and setting content: (inclused Humidity Analyzer)
Measuring related items: Oxygen concentration (vol%O2), output current value (mA), air ratio,
moisture quantity (in hot gases) (vol%H2O), mixture ratio (kg/kg),
relative humidity (%RH) and dew point (°C)
Display Items: Cell temperature (°C), thermocouple reference junction temperature (°C),
maximum/minimum/average oxygen concentration (vol%O2), maxi-mum/
minimum/average moisture quantity (vol% H2O), maximum/minimum/
average mixture ratio (kg/kg), cell e.m.f. (mV), cell internal resistance (Ω),
cell condition (in four grades), heater on-time rate (%), calibration record
(ten times), time (year/month/day, hour/minute)
Calibration Setting Items: Span gas concentration (vol%O2), zero gas concentration (vol%O2),
calibration mode (auto, semi-auto, manual), calibration type and
method (zero-span calibration, zero calibration only, span calibration
only), stabilization time (min. · sec.), calibration time (min. · sec.),
calibration interval (day/hour), starting time (year/month/day, hour/
minute)
Equipment Related Items: Measuring gas selection
Output Related Items: Analog output/output mode selection, output conditions when warmingup/maintenance/calibrating (during blow back)/abnormal, oxygen
concentration at 4 mA/20 mA (vol% O2), moisture quantity at 4 mA/20
mA (vol% H2O), mixture ratio at 4 mA/20 mA (kg/ kg), time constant.
Alarm Related Items: Oxygen concentration high alarm/high-high alarm limit values (vol%O2),
oxygen concentration low alarm/low-low alarm limit values (vol%O2),
moisture quantity high alarm/high-high alarm limit values (vol% H2O),
moisture quantity lowalarm/ low-low alarm limit values (vol% H2O),
mixture ratio high alarm/high-high alarm limit value (kg/ kg), mixture ratio
low alarm/low-low alarm limit values (kg/ kg), oxygen concentration alarm
hysteresis (vol%O2), oxygen concentration/moisture quantity/mixture
ratio alarm detection, alarm delay (seconds)
Contact Related Items: Selection of contact input 1 and 2, selection of contact output 1 to 4
(abnormal, high-high alarm, high alarm, low alarm, low-low alarm,
maintenance, calibrating, range switching, warming-up, calibration gas
pressure decrease, temperature high alarm, blow back, flameout gas
detection, calibration coefficient alarm, stabilization timeout)
Converter Output: Two points mA analog output (4 to 20 mA DC (maximum load resistance
of 550 Ω)) and one mA digital output point (HART) (minimum load
resistance of 250 Ω).
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<2. Specifications>
2-14
Oxygen analyzer; Range: Any setting between 0 to 5 through 0 to 100 vol%O2 in
1 vol%O2, or partial range is available (Maximum range value/
minimum range value 1.3 or more). For the log output, the
minimum range value is fixed at 0.1 vol%O2.
4 to 20 mA DC linear or log can be selected. Input/output
isolation provided.
Humidity analyzer; Range: any setting between 0 to 5 through 0 to 100 vol% O2, 0 to
25 through 0 to 100 vol% H2O, 0 to 0.200 through 0 to 1.000 kg/
kg or partial range is available.
For the log output, the minimum range values are fixed at 0.1
vol% O2 for the oxygen concentration, 0.1 vol% H2O for the
moisture quantity, and 0.01 kg/kg for the mixture ratio.
4 to 20 mA DC linear or log can be selected. Input/output
isolation provided.
Output damping: 0 to 255 seconds.
Hold / non-hold selection, preset value setting possible with hold.
Contact Output: Four points, contact capacity 30 V DC 3 A, 250 V AC 3 A (resistive load)
Three of the output points can be selected to either normally energized
or normally de-energized status.
Delayed functions (0 to 255 seconds) and hysteresis function (0 to 9.9
vol%O2 ,0 to 9.9 vol%H2O, 0.000 to 0.010 kg/kg) can be added to high/
low alarms.
The following functions are programmable for contact outputs.
(1) Abnormal, (2) High-high alarm, (3) High alarm, (4) Low-low alarm,
(5) Low alarm, (6) Maintenance, (7) Calibration, (8) Range switching
answer-back, (9) Warm-up, (10) Calibration gas pressure decrease
(answer-back of contact input), (11) Temperature high alarm, (12) Blow
back start, (13) Flameout gas detection (answer-back of contact input),
(14) Calibration coefficient alarm, (15) Startup power stabilization timeout
alarm
Contact output 4 is set to normally operated, and fixed error status.
Converter Input:
Thermal input one point (4 to 20 mA DC)
Contact Input: Two points, voltage-free contacts
The following functions are programmable for contact inputs:
(1) Calibration gas pressure decrease alarm, (2) Range switching
(Switched range is fixed), (3) External calibration start, (4) Process alarm
(if this signal is received, the heater power turns off), (5) Blow back start
Contact capacity:
Off-state leakage current: 3 mA or less
Self-diagnosis: Abnormal cell, abnormal cell temperature (low/high), abnormal
calibration, defective A/D converter, defective digital circuit
Calibration:
Method;
zero/span calibration
Calibration mode; automatic, semi-automatic and manual (All are operated interactively
with an LCD touchscreen). Either zero or span can be skipped.
Zero calibration gas concentration setting range:
0.3 to 100 vol%O2 (in increments of 0.01 vol%O2 in smallest units).
Span calibration gas concentration setting range:
4.5 to 100 vol%O2 (in increments of 0.01 vol%O2 in smallest units).
Use nitrogen-balanced mixed gas containing 0 to 10 vol%O2 scale of
oxygen, and 80 to 100 vol%O2 scale of oxygen for standard zero gas and
standard span gas respectively.
Calibration interval; date/time setting: maximum 255 days
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<2. Specifications>
2-15
l Model and Codes
Suffix code
Option
code
ZR402G
----------------
------
Converter
thread
-P
-G
-M
-T
Model
Display
-J
-E
-G
-F
-C
Instruction manual
—
-J
-E
-C
-A
Options
Tag plates
NAMUR NE43
compliant
Description
Separate type Zirconia Oxygen/Humidity Analyzer, Converter
---------------------
G1/2
Pg13.5
M20x1.5 mm
1/2NPT
--------------------------
Japanese
English
German
French
Chinese
----------------
Japanese
English
Chinese
------
Always -A
/HS
Set for Humidity Analyzer (*1)
/H
Hood (*3)
/SCT
/PT
Stainless steel tag plate (*2)
Printed tag plate (*2)
/C2
Failure alarm down-scale:
Output status at CPU failure and hardware error is 3.6 mA or
less (*4)
/C3
Failure alarm up-scale:
Output status at CPU failure and hardware error is 21.0 mA or
more (*4)
*1 For humidity measurements, be sure to specify /HS options.
*2 Specify either /SCT or /PT option code.
*3 Sun shield hood is still effective even if scratched.
*4 Output signal limits: 3.8 to 20.5 mA. Specify either /C2 or /C3 option code.
(Note) If AC line voltage is 125 V AC or greater, or in the EEC, the ZO21D cannot be used with the ZR402G.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<2. Specifications>
2-16
l External Dimensions
Unit: mm
1 to 6 (Panel Thickness)
2B mounting pipe
4 - Ø6 holes
for Wall mounting
100
126.5
280
EXA ZR402G
20.2
(*1/2NPT)
8
40
228
40
10
57.3
36
54.7
136.3
40
111
36 23
8-G1/2, *8-1/2NPT etc
(Wiring connection)
38 24 14 38
( for wall mounting)
*: 1/2NPT with plug
274
+2
0
4 - Ø6 holes
126.5
4-R8 to R10
or
4-C5 to C8
+2
0
190
183
Wall mounting
Panel Cut-out
With sun shield hood (option code /H)
243
±2
123
±4
155.5
±3
±3
39 ±2
55
205.5
±4
94.5
±4
±3
ZR402G
251.5
Material of HOOD : Aluminum
Accessories
Item
F2-5E.ai
Bracket for mounting
Pat.No.
Qty Description
A1113EF
1 3.15A
F9554AL 1 For pipe mounting, panel mounting or wall mounting
Screw for Bracket
F9123GF
Fuse
1
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<2. Specifications>
2.5
2-17
ZA8F Flow Setting Unit and
ZR40H Automatic Calibration Unit
2.5.1
ZA8F Flow Setting Unit
This flow setting unit is applied to the reference gas and the calibration gas in a system
configuration (System 2). Used when instrument air is provided.
This unit consists of a flowmeter and flow control valves to control the flow rates of calibration gas
and reference gas.
Standard Specifications
Flowmeter Scale:
Calibration gas; 0.1 to 1.0 l/min. Reference gas; 0.1 to 1.0 l/min.
Construction:
Dust-proof and rainproof construction
Case Material:
SPCC (Cold rolled steel sheet)
Painting:
Baked epoxy resin, Dark-green (Munsell 2.0 GY 3.1/0.5 or equivalent)
Tube Connections:
Rc1/4 or 1/4FNPT
Reference Gas pressure: Clean air supply of sample gas pressure plus approx. 50 kPaG (or
sample gas pressure plus approx. 150 kPa G when a check valve is
used). Pressure at inlet of the Flow Setting Unit.(Maximum 300 kPaG)
Air Consumption:
Approx. 1.5 l/min
Weight:
Approx. 2.3 kg
Calibration gas (zero gas, span gas) Consumption: Approx. 0.7 l/min (at calibration time only)
NOTE
Use instrument air for span calibration gas, if no instrument air is available, contact YOKOGAWA.
l Model and Codes
Suffix code
Option code
ZA8F
Model
--------
---------
Flow setting unit
Joint
-J
-A
-----------------
Rc 1/4
With 1/4 NPT adapter
---------
Style C
Style code
*C
Description
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
2-18
<2. Specifications>
l External Dimensions
180
ø6 Hole
Unit : mm
140
7
REFERENCE CHECK
REFERENCE SPAN
235.8
ZERO
2B mounting pipe
222.8
Calibration gas outlet
Span gas inlet
Reference gas outlet
Zero gas inlet
32
Piping connection port A
AIR
IN
SPAN
IN
CHECK
OUT
REF
OUT
70
ZERO
IN
Model
35
20
35
35
8
35
35
20
Instrument air inlet
Piping connection port A
ZA8F-J*C
5 - Rc1/4
ZA8F-A*C
5 - 1/4NPT
Weight : Approx. 2.3 kg
PIPNG INSIDE THE FLOW SETTING UNIT
REF
OUT
CHECK
OUT
Flowmeter
Flowmeter
AIR IN
ZERO
GAS IN
SPAN
GAS IN
Instrument air
Approx. 1.5 l/min.
Air Set
Air pressure ;
without check valve ; sample gas pressure + approx.50 kPaG
with check valve ; sample gas pressure + approx.150 kPaG
IM 11M12A01-02E
F13-14E.ai
12th Edition : Apr. 27, 2017-00
<2. Specifications>
2.5.2
2-19
ZR40H Automatic Calibration Unit
This automatic calibration unit is applied to supply specified flow of reference gas and calibration
gas during automatic calibration to the detector in a system configuration (System 3).
l Specifications
Used when auto calibration is required for the separate type and instrument air is provided.
The solenoid valves are provided as standard.
Construction: Dust-proof and rainproof construction:
NEMA4X/IP67-only for case coating solenoid valve, not flowmeter
(excluding flowmeter)
Mounting:
2-inch pipe or wall mounting, no vibration
Materials: Body; Aluminum alloy, Piping; SUS316 (JIS), SUS304 (JIS), Flowmeter;
MA (Metha acrylate resin). Bracket; SUS304 (JIS)
Finish: Polyurethane corrosion-resistance coating, mint green (Munsell
5.6BG3.3/2.9)
Piping Connection:
Refer to Model and Codes
Power Supply:
24 V DC (from ZR402G), Power consumption; Approx.1.3W
Reference Gas Pressure: Sample gas pressure plus approx. 150 kPa (690 kPa max.), (Pressure
at inlet of the Automatic Calibration Unit)
Air Consumption:
Approx. 1.5 l/min
Weight:
Approx. 3.5 kg
Ambient Temperature: -20 to +5°C, no condensation or freezing
Ambient Humidity:
0 to 95%RH
Storage Temperature: -30 to 65°C
l Model and Codes
Suffix code
Option code
ZR40H
Model
-------------
--------
Automatic calibration unit for ZR402G
Gas piping
connection
-R
-T
---------------
Rc 1/4
1/4 NPT
-----------------------------
Pipe connection (G1/2)
Pg 13.5
M20 x 1.5 mm
1/2 NPT
--------
Always -A
Wiring
connection
—
-P
-G
-M
-T
-A
Description
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<2. Specifications>
2-20
l External Dimensions
Unit: mm
Wiring inlet ; 2-G1/2,Pg13.5,M20X1.5 or 1/2NPT(Female)
2B pipe mounting example
(wiring inlet is at same position on rear)
*1 with four ISO M6 screws can wall-mount
90
54
41.2
*1
4 - Ø6.5
41.2
140
49.5
12
Connection port
116.5
71.5
Flowmeter
OCK
Terminal box
223
250
Needle valve
Setting Valve for
reference gas
Setting Valve for
calibration gas
2B mounting pipe
Zero gas inlet
Rc1/4 or 1/4 NPT(Female)
46
102
16
58 MAX
Calibration gas outlet
Rc1/4 or 1/4 NPT(Female)
42
26
26
40
Reference gas outlet
Rc1/4 or 1/4 NPT(Female)
30
47.5
25
Reference gas inlet
Rc1/4 or 1/4 NPT(Female)
F2-6_2E.ai
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
2-21
<2. Specifications>
l Piping Diagram
AC-Z
Zero
EV1
AC-S
AC-C
REF
OUT
CHECK
OUT
ZR40H Automatic
Calibration Unit
ZR402G Converter
Flowmeter
Flowmeter
*2
*2
Solenoid valve
EV1,2
Span
EV2
EV1
EV2
F35_00.ai
AIR IN
ZERO GAS IN
Instrument air Approx. 1.5 l/min
*2
Needle valve is supplied as on accessory with flowmeter
2.6
F2-6_3E.ai
ZO21S Standard Gas Unit
This is a handy unit to supply zero gas and span gas to the detector in a system configuration
based on System 1. It is used in combination with the detector only during calibration.
The ZO21S does not conform to CE marking.
l Standard Specifications
Function: Portable unit for calibration gas supply consisting of span gas (air) pump,
zero gas cylinder with sealed inlet, flow rate checker and flow rate needle
valve.
Sealed Zero Gas Cylinders (6 provided): E7050BA
Capacity:
1L
Filled pressure:
Approx. 686 kPaG (at 35°C)
Composition:
0.95 to 1.0 vol%O2 + N2 balance
Power Supply:
100, 110, 115, 200, 220, 240 V AC ±10%, 50/60 Hz
Power Consumption: Max. 5 VA
Case material:
SPCC (cold rolled steel sheet)
Paint Color: Mainframe; Munsell 2.0 GY3.1/0.5 equivalent
Cover;
Munsell 2.8 GY6.4/0.9 equivalent
Piping:
Ø6 х Ø4 mm flexible tube connection
Weight:
Approx. 3 kg
l Model and Codes
Model
Suffix code
Option code
ZO21S
-----------
-------------
Standard gas unit
Power
supply
-2
-3
-4
-5
-7
-8
-------------------------------------------------------------------------
200 V AC 50/60 Hz
220 V AC 50/60 Hz
240 V AC 50/60 Hz
100 V AC 50/60 Hz
110 V AC 50/60 Hz
115 V AC 50/60 Hz
-------------------------
Japanese version
English version
-------------
Style A
Panel
Style code
-J
-E
*A
Description
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<2. Specifications>
2-22
l External Dimensions
1600
253
228
92
Unit: mm
Flow checker
Span gas valve
Zero gas valve
Gas outlet
354
Zero gas cylinder (6 cylinder): E7050BA
2.7
F2-7E.ai
Other Equipments
2.7.1
Dust Filter for the Detector (K9471UA)
This filter is used to protect the detector sensor from a corrosive dust components or high velocity
dust in recovery boilers and cement kilns.
Sample gas flow rate is needed to be 1 m/sec or more to replace gas inside zirconia sensor.
l Standard specification
Applicable detector:
Mesh:
Material:
Weight:
Standard-type detector for general use (the sample gas flow should be
approximately perpendicular to the probe.)
30 microns
Carborundum (Filter), SUS316 (JIS)
Approx. 0.2 kg
Part No.
Description
K9471UA
Filter
K9471UX
Tool
32
Detector
Ø51
Carborundum filter (SiC)
Attach the filter unit to the tip of
the detector by screwing it clockwise.
Screw
10
Increasing of insertion length
F3-2E.ai
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
2.7.2
<2. Specifications>
2-23
Dust Guard Protector (K9471UC)
Recommended to be used when sample gas is likely to flow directly into the cell due to its flow
direction in the stack or the like, flammable dust may go into the cell, or water drops are likely to
fall and remain in the cell during downtime or the like due to the installation position.
Material:
SUS316
Weight:
Approx. 0.3 kg
Increasing of insertion length
Ø50.8
100
4-Ø6
122
F11-1E.ai
2.7.3
Ejector Assembly for High Temperature (E7046EC,
E7046EN)
This ejector assembly is used where pressure of sample gas for high temperature detector is
negative. This ejector assembly consists of an ejector, a pressure gauge assembly and a needle
valve.
l Standard Specifications
Ejector
Ejector Inlet Air Pressure: 29 to 69 kPa G
Air Consumption:
Approx. 30 to 40 l/min
Suction gas flow rate: 3 to 7 l/min
Connection:
Rc1/4, SUS304 (JIS)
Tube Connection:
Ø6 / Ø4 mm or 1/4 inch copper tube (stainless tube)
Pressure Gauge Assembly
Material in Contact with Gas: SUS316 (JIS)
Case Material:
Aluminum alloy (Paint color; black)
Scale:
0 to 100 kPaG
Connection: R1/4 or 1/4NPT, SUS304 (JIS) (with Bushing G3/8 x R1/4 or
1/4NTP)
Needle Valve
Connection:
Rc1/4 or 1/4FNPT
Material:
SUS316 (JIS)
(Note) Pipe and connections are not provided.
Part No.
Description
E7046EC
Ejector; Ø6 / Ø4 TUBE joint, Pressure gauge; R1/4,
Needle valve; Rc1/4: SUS304 (JIS)
E7046EN
Ejector; 1/4 TUBE joint, Pressure gauge; 1/4NPT(M),
Needle valve; 1/4FNPT : SUS304 (JIS)
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<2. Specifications>
2-24
< Pressure setting for the ejector assembly for high temperature use >
Pressure supply for the ejector assembly should be set so that the suction flow of the sample gas
becomes approximately 5 l/min.
To set this, proceed as follows:
(1) In Graph 4, draw a horizontal line from the 5 l/min point. on the vertical axis (Suction flow:
Qg) toward the gas pressure line to be used, to find the point of intersection.
Draw a line vertically down from the point of intersection to the axis to find the drive
pressure, P (at the ejector entrance).
(2) In Graph 1, determine Po (pressure setting) from L (the distance between the ejector and
the pressure gauge).
(3) Open the needle valve to supply air for the ejector to the pressure gauge until it indicates the
pressure setting, Po.
NOTE
Qg (the suction flow) may require change according to the conditions of use. Refer to Section
3.2.1 and Section 4.1.4 for details.
Graph explanation
(1) Graph 1 is to compensate for pressure loss in piping between the ejector and the pressure
gauge, and find Po (pressure setting).
(2) Graph 2 shows correlation between P (drive pressure) and Qa (air consumption).
(3) Graph 3 shows correlation between P (drive pressure) and Pg (suction pressure; when the
sample gas inlet of the ejector is closed).
(4) Graph 4 shows correlation between P (drive pressure) and Qg (suction flow) for each gas
pressure.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<2. Specifications>
2-25
Pressure gauge assembly
39
Pressure gauge
Ejector
Needle valve
<1>
Approx. 70
38
Instrument air inlet
<2>
Approx. 67
Full open height
Approx. 88
Ø43
<3>
Detector
R1/2
20
Tee
40
Nozzle (Note1)
Blow Rc1/4
(Note1) The connecter of ejector assembly is a
dedicated connecter with nozzle function.
Needle
valve
100
1)
5
L (m)
L
Pg
2)
Qa (l/min)
40
30
10
15
0
Pressure setting characteristics
Pg (kPa)
Ejector
Qa
Qg
P= 0.5
0
Pressure gauge
Po
Air
source
Po (kPa)
200
<1> Rc1/4 or 1/4 FNPT
<2> Ø6/Ø4 or 1/4 inch copper tube (stainless)
with ejector to connect
<3> R1/4 or 1/4 NPT
40
P (kPa)
60
80
Air consumption characteristics
4)
3)
Qg (l/min)
Gas Pressure:
8
0 kPa
-1.0
-0.5
4
0
0
Gas pressure : -15 Pa
60
40
P (kPa)
80
Suction pressure characteristics
40
60
Gas
Po (kPa)
P
(kPa)
Pg (kPa)
Qa (l/min)
Qg (l/min)
L
(m)
: Pressure setting
: Drive pressure (at the ejector entrance)
: Suction pressure
: Air consumption
: Suction flow
: Distance between the ejector and the pressure
gauge
80
P (kPa)
Suction flow characteristics
F2-9E.ai
2.7.4
Stop Valve (L9852CB, G7016XH)
This valve mounted on the calibration gas line in the system to allow for manual calibration. This
is applies to the system configuration shown for system 1 in section 1.
Standard Specifications
Connection:
Material:
Weight:
Rc 1/4 or 1/4 NPT(F)
SUS 316 (JIS)
Approx. 150 g
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<2. Specifications>
Part No.
Description
L9852CB
Joint: Rc 1/4, Material: SUS316 (JIS)
G7016XH
Joint: 1/4 NPT, Material: SUS316 (JIS)
Unit: mm
58
(Full open length)
Ø48
Rc1/4 or 1/4NPT
40
2.7.5
2-26
F15.ai
Check Valve (K9292DN, K9292DS)
This valve is mounted on the calibration gas line (directly connected to the detector).
This is applied to a system based on the (System 2 and 3) system configuration .
This valve prevents the sample gas from entering the calibration gas line. Although it functions as
a stop valve, operation is easier than a stop valve as it does not require opening/closing at each
calibration.
Screw a check valve, instead of a stop valve into the calibration gas inlet of the detector.
l Standard Specifications
Connection:
Material:
Pressure:
Weight:
Part No.
Rc1/4 or 1/4NPT(F)
SUS304 (JIS)
70 kPa G or more and 350 kPa G or less
Approx. 90 g
Description
K9292DN
Joint: Rc 1/4, Material: SUS304 (JIS)
K9292DS
Joint: 1/4 NPT, Material: SUS304 (JIS)
K9292DN : Rc 1/4(A),R 1/4(B)
K9292DS : 1/4FNPT(A),1/4NPT(Male)(B)
A
Approx. 19
Unit: mm
B
Approx. 54
F2-11E.ai
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<2. Specifications>
2.7.6
2-27
Air Set
This set is used to lower the pressure when instrument air is used as the reference and span
gases.
l Standard Specifications
G7003XF, K9473XK
Primary Pressure:
Secondary Pressure:
Connection:
Weight:
Max. 1 MPa G
0.02 to 0.2 MPa G
Rc1/4 or 1/4NPT(F) with joint adapter
Approx.1 kg
Part No.
Description
G7003XF
Joint: Rc 1/4, Material: Zinc alloy
K9473XK
Joint: 1/4 NPT(F) with adapter, Material: Zinc alloy, Adapter: SUS 316
G7004XF, K9473XG
Primary Pressure:
Secondary Pressure:
Connection:
Weight:
Max. 1 MPa G
0.02 to 0.5 MPa G
Rc1/4 or 1/4NPT(F) with joint adapter
Approx. 1 kg
Part No.
Description
G7004XF
Joint: Rc 1/4, Material: Zinc alloy
K9473XG
Joint: 1/4 NPT(F) with adapter, Material: Zinc alloy, Adapter: SUS 316
l External Dimensions
Unit : mm
View A
Panel cut dimensions
Horizontal
mounting
Vertical
mounting
ø15
22
40
+0.5
2-ø2.2 -0
40
2-ø6.5
max. 55
2-ø6 screw depth 8
Primary
Max. 210
Ø74
Panel (Horizontal mounting)
Secondary
pressure
gauge
Secondary
Panel (Vertical mounting)
A
88
G7003XF, G7004XF: Rc 1/4
K9473XK, K9473XG: 1/4NPT connector
Approx. 122
airset.ai
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
2-28
<2. Specifications>
2.7.7
Zero Gas Cylinder (G7001ZC)
The gas from this cylinder is used as the calibration zero gas and detector purge gas.
l Standard Specifications
Capacity:
Filled pressure:
Composition:
Weight:
3.4 L
9.8 to 12 MPa G
0.95 to 1.0 vol%O2 in N2
Approx. 6 kg
(Note) Export of such high pressure filled gas cylinders to most countries is prohibited or restricted.
325
485
Unit : mm
Φ140
2.7.8
F2-13E.ai
Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
This pressure reducing valve is used with the zero gas cylinders.
l Standard Specifications
Primary Pressure:
Secondary Pressure:
Connection:
Material:
Max. 14.8 MPa G
0 to 0.4 MPa G
Inlet; W22 14 threads, right hand screw
Outlet; Rc1/4 or 1/4NPT(F)
Brass body
Approx.112
Secondary pressure gauge
Primary pressure gauge
Reducing valve handle
W22 (Righthand screw)
ACH
IN
O
IH
TAK
Stop valve
* Outlet
Secondary safety valve
Primary safety valve
Approx. 59
Approx. 82
Approx. 163
Approx. 174
Part No.
* Outlet
G7013XF
Rc1/4
G7014XF
1/4 NPT female screw
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<2. Specifications>
2.7.9
2-29
Case Assembly for Calibration Gas Cylinder
(E7044KF)
This case assembly is used to store the zero gas cylinders.
l Standard Specifications
Installation:
Material:
Case Paint:
Weight:
2B pipe mounting
SPCC (Cold rolled steel sheet)
Baked epoxy resin, Jade green (Munsell 7.5 BG 4/1.5)
Approx. 10 kg with gas cylinder
(Note) Export of such high pressure filled gas cylinder to most countries is prohibited or restricted.
Pressure reducing valve
225
180
( G7013XF / G7014XF )
324
Zero gas cylinder
496
(G7001ZC)
(160)
(158.3)
2B mounting pipe
( Ø60.5 )
The oblique line is an opening portion.
(Note)The zero gas cylinder and the reducing valve are not included in the E7044KF (case assembly)
F2-15E.ai
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<2. Specifications>
2.7.10
2-30
ZR22A Heater Assembly
l Model and Codes
Style: S2
Model
Suffix code
Option code
ZR22A
---------
-----------
Heater Assembly for ZR22G
Length (*1)
-015
-040
-070
-100
-150
-200
-250
-300
---------------------------------------------------------------------------------
0.15 m
0.4 m
0.7 m
1m
1.5 m
2m
2.5 m
3m
---------------------
with Jig
None
-A
-----------
-B
-C
---------------------
Natural convention,
External connection (Instrument air)
Pressure compensated (for ZR22G S2)
Pressure compensated (for ZR22G S1)
Jig for change
-A
-N
Reference gas (*3)
Description
(*2)
*1 Suffix code of length should be selected as same as ZR22G installed.
*2 Jig part no. is K9470BX to order as a parts after purchase.
*3 Select appropriately among “-A”, “-B”, “-C” according to the reference gas supply method and style.
(Note) The heater is made of ceramic, do not drop or subject it to pressure stress.
Jig for change
Unit : mm
30
Φ 45
(K9470BX)
K9470BX
l External Dimensions
Φ 21.7
L±12
Model & Codes
L
Weight (kg)
ZR22A-015
302
Approx. 0.5
ZR22A-040
552
Approx. 0.8
ZR22A-070
852
Approx. 1.2
ZR22A-100
1152
Approx. 1.6
ZR22A-150
1652
Approx. 2.2
ZR22A-200
2152
Approx. 2.8
ZR22A-250
2652
Approx. 3.4
ZR22A-300
3152
Approx. 4.0
F2-16E.ai
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
3.
<3. Installation>
3-1
Installation
This chapter describes installation of the following equipment:
Section 3.1
General-purpose Detector (except ZR22G-015)
Section 3.2
High Temperature Detector (ZR22G-015)
Section 3.3
Converter
Section 3.4
ZA8F Flow Setting Unit
Section 3.5
ZR40H Automatic Calibration Unit
Section 3.6
Case Assembly (E7044KF)
3.1
Installation of General-purpose Detector
The following should be taken into consideration when installing the detector:
(1) Easy and safe access to the detector for checking and maintenance work.
(2) An ambient temperature of not more than 150°C, and the terminal box should not be
affected by radiant heat.
(3) A clean environment without any corrosive gases.
NOTE
A natural convection type detector (model ZR22G----C), which uses ambient air as
reference gas, requires that the ambient oxygen concentration be constant.
(4) No vibration.
(5) The sample gas satisfies the specifications described in Chapter 2.
(6) No sample gas pressure fluctuations.
3.1.1
Probe Insertion Hole
Includes those analyzers equipped with a probe supporter and probe protector.
When preparing the probe insertion hole, the following should be taken into consideration:
CAUTION
• The outside dimension of detector may vary depending on its options. Use a pipe that is
large enough for the detector. Refer to Figure 3.1 for the dimensions.
• If the detector is mounted horizontally, the calibration gas inlet and reference gas inlet
should face downwards.
• When using the detector with pressure compensation, ensure that the flange gasket does
not block the reference gas outlet on the detector flange. If the flange gasket blocks the
outlet, the detector cannot conduct pressure compensation. Where necessary, make a
notch in the flange gasket. Confirm the external dimensions of the detector in Section 2.2
before installation.
• The sensor (zirconia cell) at the tip of the detector may deteriorate due to thermal shock if
water drops are allowed to fall on it, as it is always at high temperature.
(1) Do not mount the probe with the tip higher than the probe base.
(2) If the probe length is 2.5 meters or more, the detector should be mounted vertically (no more
than a 5° tilt).
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
3-2
<3. Installation>
(3) The detector probe should be mounted at right angles to the sample gas flow or the probe
tip should point downstream.
*1
Bounds of the probe
insertion hole location
(vertical)
Flange matches
the detector size
100 mm
*1
Note
Type
Outside diameter of detector
Standard
50.8 mm in diameter (Note)
With dust filter
51 mm in diameter (Note)
With probe protector 60.5 mm in diameter (Note)
80 mm in diameter or longer (Note)
With dust protector
(horizontal)
100 mm
Four-hole flange
JIS flange
Eight-hole flange (the detector with
dust protector)
F3-1E.ai
(Note) When using the detector with pressure compensation, ensure that the flange gasket does not block the reference gas outlet on
the detector flange. If the flange gasket blocks the outlet, the detector cannot perform pressure compensation.
Where necessary, make a notch in the flange gasket. Confirm the outside dimensions of the detector in Section 2.2 before
installation.
When using the detector with ZH21B dust protector the diameter of the hole should be 80mm or larger.
Figure 3.1
3.1.2
Illustrates an example of the probe insertion hole.
Installation of the Detector
CAUTION
• The cell (sensor) at the tip of the detector is made of ceramic (zirconia). Do not drop the
detector, as impact will damage it.
• A gasket should be used between the flanges to prevent gas leakage. The gasket material
should be heatproof and corrosion-proof, suited to the characteristics of the sample gas.
The following should be taken into consideration when mounting the general use detector:
(1) Make sure that the cell mounting screws (four bolts) at the tip of the detector are not loose.
If a dust filter (see Section 2.7.1) is used, make sure it is properly attached to the detector.
Refer to Section 3.1.3 for installation of the dust filter.
(2) Where the detector is mounted horizontally, the calibration gas inlet and the reference gas
inlet should face downward.
3.1.3
Installation of the Dust Filter (K9471UA), Dust Guard
Protector (K9471UC), Probe Protector ZO21R
CAUTION
• The dust filter is used to protect the Zirconia sensor from corrosive dust or a high
concentration of dust such as in utility boilers and cement kilns. If a filter is used in
combustion systems other than these, it may have adverse effects such as response delay.
These combustion conditions should be examined carefully before using a filter.
• The dust filter requires gas flow of 1 m/sec or faster at the front surface of the filter.
When you specify option code /F1, the detector is shipped with the dust filter mounted.
Follow this procedure replace the filter on the detector . It is recommended that you read Chapter
11 prior to filter mounting, for it is necessary to be familiar with the detector’s construction,
especially the sensor assembly.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<3. Installation>
3-3
(1) Mount the dust filter assembly by putting it on the end of the detector and screw the
assembly clockwise. Put a hook pin wrench (K9471UX), Ø52 to Ø55 in diameter, into the
hole on the assembly to fasten or remove it.
Apply a heat-resistant coating (see Note 1) to the threads on the detector.
When remounting filter assembly after having once removed it from the detector, reapply the
heat-resistant coating.
Note 1:
As the detector is heated to 700°C, it is recommended to use heat-resistant coating on the threads to prevent seizing up.
Name of the heat-resistant coating material: NEVER SEEZ Nickel Special”.
32
Detector
Ø51
Carborundum filter (SiC)
Attach the filter unit to the tip of
the detector by screwing it clockwise.
Screw
10
Increasing of insertion length
F3-2E.ai
Figure 3.2
Installation of the dust filter
< Procedures for installing the dust guard protector (K9471UC)>
The ZR22G detector is shipped with the dust guard protector when the option code / F2 is
specified in case of ordering the detector. The protector should be used when preventing dusts
and water drops from lowering the detector performance is desired.
Screw the protector on the top of the detector so as to cover the top. When attaching or detaching
the protector, perform by hooking holes of its side with a hook pin wrench for Ø52 to Ø55 hole
(Pin diameter 4.5 mm: P/N K9741UX or the like) or by pass a screwdriver through the holes.
When re-attaching the protector after detaching it, apply the “NEVER SEEZ Nickel Special” to it.
Increasing of insertion length
Ø50.8
100
4-Ø6
122
F11-1.ai
Figure 3.3
The detector is used with a probe protector to support the probe (ZR22G) when the probe length
is 2.5 m or more and it is mounted horizontally.
(1) Put a gasket (provided by the user) between the flanges, and mount the probe protector in
the probe insertion hole.
(2) Make sure that the sensor assembly mounting screws (four bolts) at the tip of the detector
are not loose.
(3) Mount the detector so that the reference gas and calibration gas inlet faces downward.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<3. Installation>
Flow direction of the
sample gas
3-4
Unit: mm
Gasket (t1.5)
1050,1550,2050
Φ60.5
Notch
Probe top
Reference gas inlet
Calibration gas inlet
F3-3E.ai
Figure 3.4
Probe protector (supporting the mounting strength)
The detector is used with a probe protector to prevent the sensor from being worn by dust
particles when there is a high concentration of dust and gas flow exceeds 10 m/sec (pulverized
coal boiler or fluidized-bed furnace).
(1) Put a gasket (provided by the user) between the flanges, and mount the probe protector
in the probe insertion hole. The probe protector should be installed so that the notch is
downstream of the sample gas flow.
(2) Make sure that the sensor assembly mounting screws (four bolts) at the probe tip are not
loose.
(3) Where the detector is mounted horizontally, the reference gas and calibration gas inlet
should face downward.
CAUTION
When the probe protector is used in the ZR22G with pressure compensation (-P), instrument air
leaking from the probe protector may affect the measured value.
1050,1550,2050
Gasket (t1.5)
Unit: mm
Direction of the sample gas flow
Probe top
Mount the protector so that the notch is
the downstream side of the gas flow.
F3-4E.ai
Figure 3.5
3.1.4
Reference gas inlet
Calibration gas inlet
Mounting of detector with a probe protector
Installation of ZH21B Dust Protector
(1) Put the gasket that is provided by the user between the flanges and mount the dust
protector in the probe insertion hole.
(2) Make sure that the cell assembly mounting screws (four) at the probe tip are not loose.
(3) Mount the detector so that the calibration gas inlet and the reference gas inlet face
downward.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<3. Installation>
3-5
Unit : mm
Reference gas inlet
Calibration gas inlet
F3-2E.ai
Figure 3.6
3.2
Installation of the dust filter
Installation of High Temperature Detector
(ZR22G-015)
This detector is used with the High Temperature Probe Adapter (Model ZO21P-H) when the
temperature of sample gas exceeds 700°C, or when it is required due to maintenance spaces.
The following should be taken into consideration when installing the detector:
(1) Easy and safe access to the detector for checking and maintenance work.
(2) Ambient temperature of not more than 150°C and the terminal box should not be exposed to
radiant heat.
(3) A clean environment without any corrosive gases.
(4) No vibration.
(5) The sample gas should satisfy the specifications described in Chapter 2.
3.2.1
Usage of the High Temperature Probe Adapter
(ZO21P-H)
During analysis, the surface temperature of the probe adapter should be within the range from
more than the dew point of the sample gas and 300°C or less to prevent ejector clogging, gasket
deterioration or bolt scoring.
Where the dew point of the sample gas is not known, keep within the range of more than 200°C
to less than 300°C.
The temperature shall be measured at the probe in the probe adapter and the surface of the blind
flange at the opposite side.
When the surface temperature is not within the above range, the following measures can be
taken to change the temperature.
(1) When the furnace pressure is negative, lower the pressure setting to reduce induction flow
of the sample gas.
Refer to Section 2.7.3, Ejector Assembly for High Temperature, for the setting of induction
flow. When you reduce induction flow, ensure that the ejector inducts air when the furnace
pressure fluctuates.
(2) When the furnace pressure is positive, close the needle valve for the sample gas outlet to
reduce the exhaust gas flow. Refer to Section 4.1.4, Piping to the High Temperature Probe
Adapter.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
3-6
<3. Installation>
(3) When the probe adapter is surrounded by a heat insulator, remove the heat insulator.
Ensure that the temperature of the probe adapter does not fall below the dew point of the
gas in winter.
(4) To prevent temperature rises due to radiant heat, insert heat insulator between the wall of
the furnace and the probe adapter.
(5) To prevent temperature rises from thermal conduction, place the mounting flange as far
from the wall of the furnace as possible.
(1) When the furnace pressure is negative, raise the pressure setting to increase induction flow
of the sample gas.
Refer to Section 2.7.3, Ejector Assembly for High Temperature, for the setting of induction flow.
If there is much dust in the gas, the ejector may become clogged as induction flow increases.
(2) When the furnace pressure is positive, open the needle valve of the sample gas outlet to
increase the gas flow.
Refer to Section 4.1.4, Piping to the High Temperature Probe Adapter.
(3) Warm the probe adapter. Refer to Section 4.1.4, Piping to the High Temperature Probe
Adapter.
(4) When the surface temperature is still less than 200°C or below the dew point of the sample
gas, even if the above measures have been taken, warm the probe adapter using a heat
source such as steam.
3.2.2
Probe Insertion Hole
A high temperature detector consists of a ZR22G-015 Detector and ZO21P High Temperature
Probe Adapter. When forming the probe insertion hole, the following should be taken into
consideration:
(1) If the probe is made of silicon carbide (SiC), the probe hole should be formed so that the
probe is mounted vertically (within ± 5° tilt).
(2) In the case where the probe is made of stainless steel and the probe adapter (ZO21P-H-B)
is to be mounted horizontally, the probe hole should be formed so that the probe tip is not
higher than the probe base.
Figure 3.7 illustrates examples of the probe insertion hole.
JIS 5K 50 FF (equivalent) or
ANSI Class 150 4 RF
(equivalent) flange
100mm
A space of 52 mm or more
in diameter, and long enough
for the length of the probe
100mm
A space of 52 mm or
more in diameter, and
long enough for the length
of the probe
Never mount the probe
with its tip higher than
the base
JIS 5K 50 FF (equivalent)
or ANSI Class 150 4 RF
(equivalent) flange
The allowable margin for probe
tilt is within ± 5°.
An SiC probe shall be mounted vertically.
Horizontal mounting is used with a SUS probe.
F3-5E.ai
Figure 3.7
Examples of the probe insertion hole
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
3.2.3
<3. Installation>
3-7
Mounting of the High Temperature Detector
CAUTION
• Ceramic (zirconia) is used in the sensor (cell) portion on the detector probe tip. Care should
be taken not to drop the detector during installation.
• The same applies to a probe made of silicon carbide (SiC).
• A gasket should be used on the flange surface to prevent gas leakage. The gasket material
should be selected depending on the characteristics of the sample gas. It should be heatproof
and corrosion-proof. The parts, which should be supplied by the user, are listed in Table 3.1.
Table 3.1
Accessories for mounting high temperature probe adapter
Mounting flange specification
JIS 5K 50 FF (equivalent)
Q’ty
Note
Gasket
Parts name
1
Heatproof and corrosion-proof
Bolt (M12 by 50)
4
Nut (M12)
4
Washer (for M12)
8
ANSI Class 150 4RF (equivalent) Gasket
1
Bolt (M16 by 60)
8
Nut (M16)
8
Washer (for M16)
16
Heatproof and corrosion-proof
A high temperature detector should be mounted as follows:
(1) It is recommended to mount the detector vertically. When it is impossible due to the physical
arrangements and the detector is mounted horizontally, ensure that the probe tip be placed
no higher than the probe base.
(2) When mounting a high temperature probe adapter, be sure to insert a gasket between the
flanges to prevent gas leakage. When the furnace pressure is negative, ensure that there is
no leakage of air into the detector.
(3) When mounting the detector in a position other than vertical, the cable inlet should face
downward.
(4) When installing the detector in a low-temperature location such as in the open air, cover the
probe adapter including the ejector with a heat insulator (e.g. ceramic wool) to keep it warm
and to prevent condensation of drain on the ejector.
Wiring connection
Mount the probe so that
the probe tip is not higher
than the probe base.
Wiring connection
When using an SiC probe, mount the probe vertically
with an allowable tilt of ± 5°.
F3-6E.ai
Figure 3.8
Mounting of the High Temperature Detector
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
3.3
<3. Installation>
3-8
Installation of the ZR402G Converter
The following should be taken into consideration when installing the converter:
(1) Readability of the indicated values of oxygen concentration or messages on the converter
display. Easy and safe access to the converter for operating keys on the panel.
(2) Easy and safe access to the converter for checking and maintenance work.
(3) An ambient temperature of not more than 55°C and little change in temperature
(recommended within 15°C in a day).
(4) The normal ambient humidity (recommended between 40 to 75%RH) and without any
corrosive gases.
(5) No vibration.
(6) Near to the detector.
(7) Not in direct rays of the sun. If the sun shines on the converter, prepare the hood (/H) or
other appropriate sunshade.
n Mounting of the Converter
The converter can be mounted on a pipe (nominal JIS 50A: O.D. 60.5 mm), a wall or a panel. The
converter can be mounted at an angle to the vertical, however, it is recommended to mount it on
a vertical plane.
Mount the converter as follows.
(1) Prepare a vertical pipe of sufficient strength (nominal JIS 50A: O.D. 60.5 mm) for mounting
the converter. (Converter weighs approximately 6 kg.)
(2) Mount the converter on the pipe. Fix it firmly on the pipe in the procedure described in
Figure 3.9.
Mounting procedure
1. Put four bolts in the holes on the fitting.
2. Clamp the pipe with the fitting and a bracket,
with the four bolts passing through the bracket holes.
3. Secure the fitting and the bracket tightly to the pipe with
four washers and nuts.
A pipe to be mounted (nominal JIS 50A)
Bracket
*Fitting
*Washer
Nut
*Bolt
Bolt
* Note: These fittings are attached to the converter when it is delivered.
Figure 3.9
Washer
F3-7E.ai
Pipe Mounting
(1) Drill mounting holes through the wall as shown in Figure 3.10.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
3-9
<3. Installation>
Unit: mm
Four holes 6 mm in diameter
for M5 screws
126.5
190
Figure 3.10
F3-8E.ai
Mounting holes
(2) Mount the converter. Secure the converter on the wall using four screws.
Note: For wall mounting, the bracket and bolts are not used.
F3-9E.ai
Figure 3.11
Wall Mounting
(1) Cut out the panel according to Figure 3.12.
Unit: mm
+2
274 0
Washer
Bolt
+2
183 0
Figure 3.12
F3-10E.ai
Panel cutout sizes
Fitting
Clamp screw
Figure 3.13
F3-11E.ai
Panel mounting
(2) Remove the fitting from the converter by loosening the four screws.
(3) Insert the converter case into the cutout hole of the panel.
(4) Attach the mounting fitting which is once removed in step (2) again to the converter.
(5) Firmly fix the converter to the panel. Fully tighten the two clamp screws to hold the panel
with the fitting.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
3.4
<3. Installation>
3-10
Installation of ZA8F Flow Setting Unit
The following should be taken into consideration:
(1) Easy access to the unit for checking and maintenance work.
(2) Near to the detector or the converter
(3) No corrosive gas.
(4) An ambient temperature of not more than 55°C and little changes of temperature.
(5) No vibration.
(6) Little exposure to rays of the sun or rain.
n Mounting of ZA8F Flow Setting Unit
The flow setting unit can be mounted either on a pipe (nominal JIS 50A) or on a wall. It should be
positioned vertically so that the flowmeter works correctly.
(1) Prepare a vertical pipe of sufficient strength (nominal JIS 50A: O.D. 60.5 mm) for mounting
the flow setting unit. (The unit weighs approximately 2 to 3.5 kg.)
(2) Mount the flow setting unit on the pipe by tightening the nuts with the U-bolt so that the metal
fitting is firmly attached to the pipe.
F3.12E.ai
Figure 3.14
Pipe Mounting
(1) Make a hole in the wall as illustrated in Figure 3.15.
(2) Mount the flow setting unit. Remove the pipe mounting parts from the mount fittings of the
flow setting unit and attach the unit securely on the wall with four screws.
Unit : mm
223
140
F3-13E.ai
4 - Φ6.5 hole, or M6 screw
Figure 3.15
Mounting holes
F3-14E.ai
Figure 3.16
Wall mounting
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
3.5
<3. Installation>
3-11
Installation of ZR40H Automatic Calibration
Unit
The following should be taken into consideration:
(1) Easy access to the unit for checking and maintenance work.
(2) Near to the detector or the converter
(3) No corrosive gas.
(4) An ambient temperature of not more than 55°C and little change of temperature.
(5) No vibration.
(6) Little exposure to rays of the sun or rain.
n Mounting of ZR40H Automatic Calibration Unit
The automatic calibration unit can be mounted either on a pipe (nominal JIS 50A ) or on a wall.
It should be positioned vertically so that the flowmeter works correctly.
(1) Prepare a vertical pipe of sufficient strength (nominal JIS 50A: O.D. 60.5 mm) for mounting
of automatic calibration unit. (The unit weights approximately 3.5 kg.)
(2) Mount the automatic calibration unit on the pipe by tightening the nuts with the U-bolt so that
the metal fitting is firmly attached to the pipe.
Figure 3.17
Pipe Mounting
(1) Make a hole in the wall as illustrated in Figure 3.18.
Unit : mm
223
140
F3-13E.ai
4 - Φ6.5 hole, or M6 screw
Figure 3.18
Mounting holes
(2) Mount the automatic calibration unit. Remove the U-bolt from the automatic calibration unit
and attach the unit on the wall with four screws. When setting it with M5 bolts, use washers.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<3. Installation>
3-12
4-Φ6.5
F3-16_1E.ai
Figure 3.19
3.6
Wall Mounting
Installation of the Case Assembly (E7044KF)
The case assembly is used to store the G7001ZC zero gas cylinders.
The following should be taken into consideration:
(1) Easy access for cylinder replacement
(2) Easy access for checking
(3) Near to the detector or converter as well as the flow setting unit.
(4) The temperature of the case should not exceed 40°C due to rays of the sun or radiated heat.
(5) No vibration
n Mounting
Mount case assembly on a pipe (nominal JIS 50A) as follows:
(1) Prepare a vertical pipe of sufficient strength (nominal JIS 50A: O.D. 60.5 mm) for mounting
the case assembly. (The sum of the case assembly and the calibration gas cylinder weighs
approximately 4.2 kg.)
(2) Mount the case assembly on the pipe by tightening the nuts with the U-bolt so that the metal
fitting is firmly attached to the pipe.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<3. Installation>
3-13
A pipe to be mounted
(nominal JIS 50A : O.D. 60.5 mm)
F3-16E.ai
Figure 3.20
3.7
Pipe Mounting
Insulation Resistance Test
Even if the testing voltage is not so great that it causes dielectric breakdown, testing may cause
deterioration in insulation and a possible safety hazard. Therefore, conduct this test only when it
is necessary.
The applied voltage for this test shall be 500 V DC or less. The voltage shall be applied for as
short a time as practicable to confirm that insulation resistance is 20 MΩ or more.
Remove wiring from the converter and the detector.
(1) Remove the jumper plate located between terminal G and the protective grounding terminal.
(2) Connect crossover wiring between L and N.
(3) Connect an insulation resistance tester (with its power OFF). Connect (+) terminal to the
crossover wiring, and (-) terminal to ground.
(4) Turn the insulation resistance tester ON and measure the insulation resistance.
(5) After testing, remove the tester and connect a 100 kΩ resistance between the crossover
wiring and ground, to discharge for over 1 min. During discharge, do not touch the terminal.
(6) Testing between the heater terminal and ground, contact output terminal and ground, analog
output/input terminal and ground can be conducted in the same manner.
(7) Although contact input terminals are isolated, insulation resistance test cannot be conducted
because the breakdown voltage of the surge-preventing arrester between the terminal and
ground is low.
(8) After conducting all the tests, replace the jumper plate as it was.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
3-14
<3. Installation>
Insulation
resistance tester
-
ZR402G Converter
ZR22G Detector
+
crossover wiring
1
FG
12
FG
2
3
4
5
6
7
8
9
10
11
AO1 AO1 AO2 AO2 CJ
CJ
TC
TC CELL CELL
(+)
(-)
(+)
(-)
(+)
(-)
(+)
(-)
(+)
(-)
13
14
15
16
17
18
19
20
21
22
DI-1 DI-2 DI-C
AI
AI
AC- AC- AC- FG
FG
(+)
(-) ZERO SPAN COM
CELL CELL
(+)
(-)
TC
(+)
TC
(-)
CJ
(+)
CJ
(-)
Contact
input 2
Solenoid valve
Span gas for automatic
calibration
Contact
input 1
Solenoid valve for
automatic calibration
Zero gas
Automatic Calibration Unit
23
24
25
26
27
28
29
30
31
32
DO-1 DO-1 DO-2 DO-2 DO-3 DO-3 DO-4 DO-4 HTR HTR
33
L
34
N
35
G
36
HTR HTR
Remove
jumper plate
Remove wiring to connect external circuits
and Connect crossover wiring
crossover wiring
Insulation
resistance tester
-
+
F3-17E.ai
Insulation
resistance tester
-
+
Figure 3.21
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<4. Piping>
4-1
4. Piping
This chapter describes piping procedures based on three typical system configurations for EXAxt
ZR Separate type Zirconia Oxygen/Humidity Analyzer.
• Ensure that each check valve, stop valve and joint used for piping do not allow leakage.
Especially, if there is any leakage of the calibration gas from pipes and joints, it may cause
clogging of the pipes or incorrect calibration.
• Be sure to conduct leakage test after piping.
• Basically, apply instrument air (dehumidified by cooling to the dew point -20°C or lower, and
removing any dust, oil mist and the like) for the reference gas.
• When the instrument uses natural convection for reference gas, ambient air near the
detector is used for reference gas; therefore the accuracy of analysis will be affected by
ambient humidity changes or the like. If more accurate analysis is necessary, use instrument
air (dehumidified to the dew point -20°C or lower, and removing any dust, oil mist and the
like) for reference gas.
Stable analyzing can be conducted when using instrument air.
4.1
Piping for System 1
The piping in System 1 is illustrated in Figure 4.1.
ZR402G Converter
ZR22G Separate type
Zirconia Oxygen/Humidity Analyzer, Detector
EXA ZR402G
Stop valve
Signal
(6-core shield cable)
~
Heater (2-core)
100 to
240 V AC
Contact input
Analog output, Contact output
Digital output (HART)
Calibration gas
~
100/110/115/200/220/240 V AC
ZO21S Standard gas unit
F4-1E.ai
Figure 4.1
Piping in System 1
CAUTION
• The stop valve should be connected directly to the detector. If any piping is present between
the detector and the stop valve, water may condense in the pipe, which may cause damage
to the sensor by rapid cooling when the calibration gas is introduced.
The stop valve should be closed except while the calibration gas is being introduced.
• If a high temperature detector is used (the sample gas temperature is 700°C or higher),
piping for the reference gas is required. In other cases, piping is required if the air around
the detector is not clean.
• The reference gas should have an oxygen concentration identical to that of fresh air (21%).
• When a high temperature detector is used, the sample gas is vented into the surrounding
air.
Therefore, the oxygen concentration required may not be obtained unless an exhaust pipe
is installed.
Piping in System 1 is as follows:
• Connect a stop valve to the nipple at the calibration gas inlet of the detector. Then mount a
joint for a 6 mm (O.D.) x 4 mm (I.D.) soft tube at the stop valve connection hole of the inlet
side (see Section 4.1.2). The tube is to be connected to this joint only during calibration.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
4-2
<4. Piping>
• It is recommended to use ZH21B dust protector to protect the probe output from dust
agitation (i.e., to prevent combustible materials from entering the probe cell) where humidity
measurements are made under dusty or combustible, such as paper dust, environment.
• If a high temperature detector is used and no piping can be installed for the reference gas,
place piping in the exhaust hole for the sample gas on the high temperature probe adapter
so that the sample gas is carried away from the vicinity of the detector (see Section 4.1.4,
Figure 4.6).
• If a high temperature detector is used and the sample gas pressure is negative, connect
an ejector assembly to the sample gas exhaust hole of the high temperature probe adapter
(see Section 4.1.4, Figure 4.3).
• If a high temperature detector is used and the pressure of the sample gas is 0.49 kPa or
higher, it is recommended that a needle valve (throttle) be used in the sample gas exhaust
of the high temperature probe adapter (see Section 4.1.4, Figure 4.4).
CAUTION
This is for lowering the sample gas temperature below 700°C. If the gas temperature is high and
the pressure is also significantly high, the sample gas temperature may not fall to below 700°C
before it reaches the detector.
On the other hand, if the sample gas temperature is lowered too much, condensation may be
produced in the High Temperature Probe Adapter. During wintertime, it is recommended that the
High Temperature Probe Adapter be protected with an insulating material to avoid condensation
forming (see Section 4.1.4, Figure 4.5).
For the usage of the High Temperature Probe Adapter, refer to Section 3.2.1.
4.1.1
Parts Required for Piping in System 1
Check that the parts listed in Table 4.1 are ready.
Table 4.1
Detector
General use
detector
Piping location
Calibration gas inlet
Parts Name
Note
Stop valve
Recommended by YOKOGAWA
(L9852CB or G7016XH)
Nipple *
Rc 1/4 or 1/4 NPT
General parts
Joint for tube connection Rc 1/4 (1/4 NPT)
for a Ø6 x Ø4 mm soft tube
High temperature
detector
(0.15 m)
General parts
Reference gas inlet
(sealed up)
(when piping is required, refer to section 4.1.3)
Calibration gas inlet
Stop valve
Recommended by YOKOGAWA
(L9852CB or G7016XH)
Nipple *
Rc 1/4 or 1/4 NPT
General parts
Joint for tube connection Rc 1/4 (1/4 NPT) for a Ø6 x Ø4 mm soft tube
Reference gas inlet
(sealed up)
(when piping is required, refer to section 4.1.3)
Sample gas outlet
Ejector assembly *
Recommended by YOKOGAWA
(E7046EC or E7046EN)
T-shaped joint of the
same diameter *
R1/4 or 1/4 NPT
General parts
Needle valve *
Rc1/4 or 1/4 NPT
General parts
Nipple of other diameter
*
R 1/2 to R 1/4 or
R 1/2 to 1/4 NPT
General parts
Note: Parts with marking * are used when required.
General parts can be found on the local market.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
4.1.2
<4. Piping>
4-3
Connection to the Calibration Gas Inlet
When carrying out calibration, connect the piping (6(O.D) ~4(I.D.) mm tube) from the standard
gas unit to the calibration gas inlet of the detector. First, mount a stop valve (of a quality specified
by YOKOGAWA) on a nipple (found on the local market) as illustrated in Figure 4.2, and mount a
joint (also found on the local market) at the stop valve tip. (The stop valve may be mounted on the
detector prior to shipping the detector.)
Note: Mount the stop valve close to the detector.
Stop valve
Nipple
Tube connection joint
F4-2E.ai
Figure 4.2
4.1.3
Connection to the calibration gas inlet
Connection to the Reference Gas Inlet
• Normally, no piping is required for the reference gas inlet when the equipment uses natural
convection for the reference gas (models ZR22G----C). Leave the plug as it is . If the
air around the detector is polluted and the necessary oxygen concentration (21 vol%O2)
cannot be obtained, prepare piping the same as which described in Section 4.2, System 2.
• When the equipment uses instrument air for the reference gas, piping is required as
described in Section 4.2, System 2 (models ZR22G----E or P).
4.1.4
Piping to the High Temperature Probe Adapter
• The sample gas should be at a temperature below 700°C before reaching the detector
sensor. If the gas is under negative pressure, it should be fed to the detector by suction.
• For usage of the probe adapter when using high temperature detector, refer to Section 3.2.1.
• If the sample gas is under negative pressure, connect the ejector assembly (E7046EC/
E7046EN) as illustrated in Figure 4.3. Mount the pressure gauge as close as possible to
the ejector assembly. However, if the ambient temperature is too high, mount the gauge in a
location with a temperature below 40°C.
Ejector assembly for
high temperature use
Detector
Adapter for
high temperature probe
F4-3E.ai
Figure 4.3
Mounting the ejector assembly
If the temperature of the sample gas exceeds the specified value and its pressure exceeds 0.49
kPa, the sample gas temperature may not be below 700°C at the detector.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<4. Piping>
4-4
In such a case, connect a needle valve (found on the local market) through a nipple (also found
on the local market) to the probe adapter sample gas exhaust (Rc 1/2) so that the sample gas
exhaust volume is restricted.
Reducing nipple
(R1/2-R1/4 or R1/2-1/4NTP)
Needle valve
F4-4E.ai
Figure 4.4
Mounting the needle valve for restricting exhaust flow of the sample gas
In cases where condensation is likely to occur in the probe adapter when the sample gas is
cooled, protect the probe adapter with an insulating material as illustrated in Figure 4.5.
Cover flange
Sample gas outlet
Adapter for high temperature probe
Detector
Insulating material
Probe
F4-5E.ai
Figure 4.5
Preventing to condensation
If the sample gas is to be vented at a distance from the detector because no reference gas piping
can be provided, an exhaust pipe should be installed as illustrated in Figure 4.6.
In addition, the exhaust pipe shall be kept warm to protect against condensation.
Exhaust pipe
Rc1/2
F4-6E.ai
Figure 4.6
Use thick piping for exhaust.
Horizontal part should be kept to the
minimum to maintain piping strength.
Exhaust pipe
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
4.2
<4. Piping>
4-5
Piping for System 2
Piping in System 2 is illustrated in Figure 4.7.
ZR22G Separate type
Zirconia Oxygen/Humidity Analyzer, Detector
ZR402G Converter
EXA ZR402G
Stop valve
or Check valve
Signal
(6-core shield cable)
to
~ 100
240 V AC
Heater (2-core)
Reference
gas
Flowmeter
Needle
valve
Contact input
Analog output, Contact input
Digital output (HART)
Air Set
Calibration
gas
Instrument air
Span gas
(Same as Zero gas Calibration)
ZA8F Flow Setting Unit
Pressure
reducing
valve
Calibration gas
unit case
Figure 4.7
Zero gas cylinder
F4-7E.ai
Piping for System 2
System 2 illustrated in Figure 4.7 requires piping as follows:
• Connect a stop valve or check valve the nipple at the calibration gas inlet of the detector.
• It is recommended to use ZH21B dust protector to protect the probe output from dust
agitation (i.e., to prevent combustible materials from entering the probe cell) where humidity
measurements are made under dusty or combustible environment.
• If a high temperature detector is used and the sample gas pressure is negative, connect
an ejector assembly to the sample gas exhaust hole of the high temperature probe adapter
(see Section 4.1.4, Figure 4.3).
• If a high temperature detector is used and the pressure of the sample gas is 0.49 kPa or
higher, it is recommended that a needle valve (throttle) be used in the sample gas exhaust
of the high temperature probe adapter (see Section 4.1.4, Figure 4.4).
CAUTION
This is for lowering the sample gas temperature below 700°C. If the gas temperature is high and
the pressure is also significantly high, the sampled gas temperature may not reduced below
700°C when reaching the detector.
On the other hand, if the sample gas temperature is lowered too much, condensation may be
produced in the high temperature probe adapter. During wintertime, it is recommended that the
high temperature probe adapter be protected with an insulating material to prevent condensation
(see Section 4.1.4, Figure 4.5).
For the usage of the high temperature probe adapter, refer to Section 3.2.1.
• If the dust sticking to the interior of the high temperature probe adapter is to be eliminated by
blow back while using the high temperature detector, the air feed for blow back should also
be taken into consideration.
CAUTION
The probe is easily clogged if too much dust is contained in the sample gas such as in a utility
boiler or cement kiln. To get rid of the dust with compressed air, the piping from the air source is
connected only during cleaning. Blow back piping can be installed for dust cleaning as illustrated
in Section 4.3.1.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
4-6
<4. Piping>
4.2.1
Piping Parts for System 2
Check that the parts listed in Table 4.2 are ready.
Table 4.2
Detector
General use
detector
Piping Parts
Piping location
Parts Name
Note
Calibration gas
inlet
Stop valve or check valve
Recommended by YOKOGAWA (L9852CB
or G7016XH) Provided by YOKOGAWA
(K9292DN or K9292DS)
Nipple *
Rc 1/4 or 1/4 NPT
Zero gas cylinder
User´s scope
Pressure reducing valve
Recommended by YOKOGAWA
(G7013XF or G7014XF)
Joint for tube connection
Rc 1/4 or 1/4 NPT
Air set
Recommended by YOKOGAWA
(G7003XF/K9473XK or G7004XF/K9473XG)
Joint for tube connection
Rc 1/4 or 1/4 NPT
Stop valve or check valve
Recommended by YOKOGAWA (L9852CB
or G7016XH) Provided by YOKOGAWA
(K9292DN or K9292DS)
Nipple *
Rc 1/4 or 1/4 NPT
Zero gas cylinder
User´s scope
Pressure reducing valve
Recommended by YOKOGAWA
(G7013XF or G7014XF)
Joint for tube connection
Rc 1/8 or 1/8 NPT
Air set
Recommended by YOKOGAWA
(G7003XF/K9473XK or G7004XF/K9473XG)
Joint for tube connection
Rc 1/4 or 1/4 NPT
Ejector assembly *
Recommended by YOKOGAWA (E7046EC
or E7046EN)
T-shaped joint of the same
diameter *
R 1/4 or 1/4 NPT
General parts
Needle valve *
Rc 1/4 or 1/4 NPT
General parts
Nipple of other diameter *
R 1/2 to R 1/4
or R 1/2 to 1/4 NPT
General parts
Reference gas
inlet
High temperature
Detector
(0.15 m)
Calibration gas
inlet
Reference gas
inlet
Sample gas
outlet
General parts
General parts
General parts
General parts
General parts
Note: Parts with marking * are used when required.
General parts can be found on the local market.
4.2.2
Piping for the Calibration Gas
This piping is to be installed between the zero gas cylinder and the ZA8F flow setting unit, and
between the ZA8F flow setting unit and the ZR22G detector.
The cylinder should be placed in a case assemble E7044KF or the like to avoid any direct
sunlight or radiant heat so that the gas cylinder temperature does not exceed 40°C.
Mount a reducing valve (specified by YOKOGAWA) on the cylinder.
Mount a check valve or stop valve (specified by YOKOGAWA) on the nipple (found on the local
market) at the calibration gas inlet of the detector as illustrated in Figure 4.8.
(The check valve or the stop valve may have been mounted on the detector when shipped.)
Connect the flow setting unit and the detector to a stainless steel pipe 6 mm (O.D.) x 4 mm or
larger (I.D.) (or nominal size 1/4 inch).
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<4. Piping>
4-7
Stop valve or Check valve
Piping for the calibration gas, 6 mm (O.D.)
by 4 mm (I.D.) Stainless steel pipe
Piping for the reference gas, 6 mm (O.D.)
by 4 mm (I.D.)Stainless steel pipe
F4-8E.ai
Figure 4.8
4.2.3
Piping for the Calibration Gas Inlet
Piping for the Reference Gas
Reference gas piping is required between the air source (instrument air) and the flow setting unit,
and between the flow setting unit and the detector.
Insert the air set next to the flow setting unit in the piping between the air source and the flow
setting unit.
Use a 6 mm (O.D.) x 4 mm or larger (I.D.) (or nominal size 1/4 inch) stainless steel pipe between
the flow setting unit and the detector.
4.2.4
Piping to the High Temperature Probe Adapter
Refer to Section 4.1.4 for the piping.
4.3
Piping for System 3
Piping in System 3 is illustrated in Figure 4.10. In System 3, calibration is automated;
however, the piping is basically the same as that of System 2. Refer to Section 4.2.
Adjust secondary pressure of both the air set and the zero gas reducing valve so that these
two pressures are approximately the same. The flow rate of zero and span gases (normally
instrument air) are set by a single needle valve.
After installation and wiring, check the calibration contact output (see Sec. 7.11.2), and adjust
zero gas reducing valve and calibration gas needle valve so that zero gas flow is within the
permitted range. Next check span gas calibration contact output and adjust air set so that span
gas flow is within the permitted range.
REF
OUT
CHECK
OUT
EV1
Flowmeter
Flowmeter
*2
*2
Solenoid valve
EV1, EV2
EV2
AIR IN
ZERO GAS IN
Instrument air Approx.
1.5 l/min.
*2: Needle valve is supplied as accessory with flowmeter
Figure 4.9
F2-6-3E.ai
Piping inside in System 3
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<4. Piping>
4-8
When blow back function is used by contact input to the ZR402G converter, install blow back
piping as illustrated in Section 4.3.1.
Note: Blow back function means the function to get rid of dust inside a probe in a high temperature probe adapter by using compressed
air, when a high temperature detector is used.
It is recommended to use ZH21B dust protector to protect the probe output from dust
agitation (i.e., to prevent combustible materials from entering the probe cell) where humidity
measurements are made under dusty or combustible environment.
ZR402G Converter
ZR22G Separate type
Zirconia Oxygen/Humidity Analyzer, Detector
EXA ZR402G
Signal
(6-core shield cable)
Check valve
~
Heater (2-core cable)
Flowmeter
Reference gas
Needle
valve
100 to 240 V AC
Contact input
Analog output, contact output
Digital output (HART)
Air Set
Calibration gas
Instrument air
ZR40H
Automatic Calibration Unit
Pressure
reducing
valve
Zero gas cylinder
Calibration gas
unit case
Figure 4.10
F4-9E.ai
Piping for System 3
Blow Back Piping
This piping is required when the blow back function is carried out. The piping described below
provides automatic blow back operation when the “ Blow back start “ command is entered to the
converter.
Solenoid valve
Pressure reducing valve
Pipe junction
Blow pipe
F4-10E.ai
Figure 4.11
Blow back Piping
The following parts are required for blow back piping.
• Flange (to be prepared as illustrated in Figure 4.12.)
• Blow pipe (to be prepared as illustrated in Figure 4.12.)
• Two-way solenoid valve: “ Open “ when electric current is on. (Found on the local market) .
• Air set (recommended by YOKOGAWA, G7003XF / K9473XK or G7004XF / K9473XG)
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<4. Piping>
4-9
Manufacture the blow pipe as illustrated in Figure 4.12, and mount it on the high temperature
probe adapter.
Unit : mm
Φ
90
4-Φ9
Stainless steel flange
8 ~ 10
Rc1/4
Φ73
Welded
Approximately
200
8 (O.D.) by 6 (I.D.) Stainless steel pipe
F4-11E.ai
Figure 4.12
4.4
Manufacturing Blow pipe and Flange
Piping for the Detector with Pressure
Compensation
The ZR22G----P Detector with Pressure Compensation may be used in System 2 and
System 3. However, it cannot use piping for high temperature probe adapter or blow back piping.
Use this detector whenever the furnace pressure exceeds 5 kPa (see Note). Even if the furnace
pressure is high, the detector can measure by adjusting pressure of the detector to the furnace
pressure using instrument air. The inside pressure of the probe will be kept the same as the
furnace pressure by feeding instrument air at higher pressure than that in the furnace.
NOTE
Sample gas pressure should not vary rapidly and widely.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<4. Piping>
Rc1/4 or 1/4NPT
Reference gas inlet
156
87
4-10
*1
2-G1/2, 2-1/2NPT etc.
Cable connection port
48
25
*1
*1
Reference gas outlet
Flange
PIPING
*1
PIPING
*1
Rc1/4 or 1/4NPT
Calibration gas inlet
Stop Valve
Flange
F4-12E.ai
*1 Dimensions may vary depending on the detector type. Refer to Section 2.2
Ensure that the flange gasket does not block the reference gas outlet.
Where necessary, make a notch on the flange
Figure 4.13
Detector with Pressure Compensation
Valve operation
(1) For safety, stop the furnace that the detector is to be installed in. If furnace internal pressure
is high, this is especially dangerous.
(2) Before starting instrument air flow, completely shut the stop valve in front of the reference
gas outlet.
(3) Check that the reference gas outlet is not blocked by a flange gasket or the like.
(4) Set the instrument air pressure higher than furnace internal pressure.
(5) Completely open the stop valve in front of the reference gas outlet and, after turning on
instrument air flow, start furnace operation. As furnace internal pressure rises, confirm
that instrument air continues to flow and adjust the valve or increase supply pressure if
necessary.
(6) After furnace internal pressure stabilizes, adjust flow.
(7) If furnace operation is stopped, stop instrument air flow and completely shut the stop valve
in front of the reference gas outlet. You may leave reference gas flowing if you wish.
CAUTION
• As far as possible do not stop the instrument air flow, to prevent the sample gas from
entering the detector and damaging the zirconia cell.
• Connect the stop valve, which is at the calibration gas inlet, directly to the detector.
If there is piping between the detector and the valve, condensation may damage the sensor
by rapid cooling when calibration gas is introduced.
Figure 4.14 illustrates an example of System 2 using a detector with pressure compensation.
Supply air pressure (flow) may vary depending on the furnace pressure. It is recommended to
use a flowmeter and an air set that is suitable for the furnace pressure.
NOTE
When using the ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit, please note that
the supplying airflow (pressure) will vary depending on the furnace pressure.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
ZR402G Converter
ZR22G Separate type
Zirconia Oxygen Analyzer, Detector
Stop valve
or
Check valve
EXA ZR402G
Signal
(6-core shield cable)
~
Heater (2 core cable)
Reference
gas
Flowmeter
100 to
240 V AC
Contact input
Analog output, Contact output
Digital output (HART)
Needle
valve
Air Set
Calibration
gas
Instrument air
Span gas
(Same as Zero gas calibration unit)
ZA8F Flow Setting Unit
Pressure
reducing
valve
Zero gas cylinder
Calibration gas
unit case
Figure 4.14
4.4.1
4-11
<4. Piping>
F4-13E.ai
System 2 using a detector with pressure compensation
Piping Parts for a System using Detector with
Pressure Compensation
Check that the parts listed in Table 4.3 are ready.
Table 4.3
Detector
Piping Parts
Piping location
Detector with Calibration gas inlet
pressure
compensation
Parts Name
Note
Stop valve or check
valve
Recommended by YOKOGAWA (L9852CB
or G7016XH) Provided by YOKOGAWA
(K9292DN or K9292DS)
Nipple *
Rc 1/4 or 1/4 NPT
Zero gas cylinder
User´s scope
General parts
Pressure reducing valve Recommended by YOKOGAWA
(G7013XF or G7014XF)
Joint for tube connection Rc 1/4 or 1/4 NPT
Reference gas inlet
Air set
General parts
Recommended by YOKOGAWA
(G7003XF/K9473XK or G7004XF/K9473XG)
Joint for tube connection Rc 1/4 or 1/4 NPT
General parts
Note: Parts with marking * are used when required.
General parts can be found on the local market.
4.4.2
Piping for the Calibration Gas
Calibration gas piping is basically identical to that of System 2. See Section 4.2.2.
4.4.3
Piping for the Reference Gas
Reference gas piping is basically identical to that of for System 2. See Section 4.2.3.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
Blank Page
<5. Wiring>
5-1
5. Wiring
In this Chapter, the wiring necessary for connection to the EXAxt ZR Separate type Zirconia
Oxygen/Humidity Analyzer is described.
5.1
General
WARNING
NEVER supply current to the converter or any other device constituting a power circuit in
combination with the converter, until all wiring is completed.
CAUTION
This product complies with CE marking.
Where compliance with CE marking is necessary, the following wiring procedure is necessary.
• Install an external switch or circuit breaker to the power supply of the converter.
• Use an external switch or circuit breaker rated 5 A and conforming with IEC 947-1 or IEC
947-3.
• It is recommended that the external switch or circuit breaker be mounted in the same room
as the equipment.
• The external switch or circuit breaker should be installed within the reach of the operator,
and marked as the power supply switch of this equipment.
Wiring procedure
Wiring should be performed according to the following procedure:
1. Be sure to connect the shield line to FG terminal of the converter.
2. The outer sheath of the signal line should be stripped to a length of 50 mm or less.
The most outer sheath of the power cable should be stripped to a length of 20 mm or less.
3. Signals may be affected by noise if signal lines, power cable and heater cable are located in
the same conduit. When using conduit, signal lines should be installed in a separate conduit
than power and heater cables.
4. Install metal blind plug(s) in unused cable connection gland(s) of the converter.
5. Metal conduit should be grounded.
6. The following cables are used for wiring:
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<5. Wiring>
Table 5.1
Cable specifications
Terminal name of converter
Name
CELL+, CELLHTR TC+, HTR TCCJ+, CJ-
Detector signal
HEATER
Detector heater
L, N
Power supply
AO-1+, AO-1-, AO-2+, AO-2-
Analog output
DO-1, DO-2, DO-3, DO-4
Contact output
Need for shields
Number of cores
О
6
2
2 or 3 *
О
2 or 4
2 to 8
AC-Z, AC-S, AC-C
Automatic Calibration unit
3
DI-1, DI-2, DI-C
Contact input
3
AI+, AI-
Temperature input
Note *:
5-2
О
2
When the case is used for protective grounding, use a 2-core cable.
WARNING
Cables that withstand temperatures of at least 80°C should be used for wiring.
CAUTION
• Select suitable cable O.D. to match the cable gland size.
• Protective grounding should be connected in ways equivalent to JIS D style (Class 3)
grounding (the grounding resistance is 100 Ω or less).
• Special consideration of cable length should be taken for the HART communication, For the
detail, refer to Section 1.1.2 of the IM 11M12A01-51E “Communication Line Requirements”.
5.1.1
Terminals for the External Wiring in the Converter
Open the front door and remove the terminal covering plate to gain access to the converter
external wiring terminals (see Figure 5.1).
CAUTION
After wiring necessary cable to the converter terminals, be sure to fix the terminal covering plate
with two screws again.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<5. Wiring>
Front door
Terminals are visible
when the terminal cover is removed.
Cable inlet
Figure 5.1
5-3
F5-2E.ai
Terminals for external wiring in the converter
5.1.2 Wiring
Connect the following wiring to the converter. It requires a maximum of seven wiring connections
as shown below.
(1) Detector output (connects the converter with the detector.)
(2) Detector heater power (connects the converter with the detector.)
(3) Analog output signal
(4) Power and ground
(5) Contact output
(6) Operation of the solenoid valve of automatic calibration unit
(7) Contact input
(8) Temperature input
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
ZR22G Separate type
Zirconia Oxygen/Humidity Analyzer, Detector
ZR402G Separate type
Zirconia Oxygen/Humidity Analyzer, Converter
Analog
Analog
output 1
4-20 mA DC output 2
Digital output 4-20 mA DC
1
FG
12
FG
2
3
4
5
6
7
8
9
10
11
AO-1 AO-1 AO-2 AO-2 CJ
CJ
TC
TC CELL CELL
(+)
(-)
(+)
(-)
(+)
(-)
(+)
(-)
(+)
(-)
13
14
15
16
17
18
19
20
21
22
DI-1 DI-2 DI-C
AI
AI
AC- AC- AC- FG
FG
(+)
(-) ZERO SPAN COM
Contact
input 2
CELL CELL
(+)
(-)
TC
(+)
TC
(-)
CJ
(+)
CJ
(-)
Span gas Solenoid valve
for automatic calibration
Contact input 1
*1
5-4
<5. Wiring>
Solenoid valve
for automatic calibration
Zero gas
Automatic Calibration Unit
ZR40H
23
24
25
26
27
28
29
30
31
32
33
34
DO-1 DO-1 DO-2 DO-2 DO-3 DO-3 DO-4 DO-4 HTR HTR
L
N
Temperature transmitter input
4 to 20mA
(Only for Humidity Analyzer)
Contact
output 1
Contact
output 2
Contact
output 3
Contact
output 4
35
G
36
FG
HTR HTR
*1
100-240V AC
50/60 Hz
*1
F5-3E.ai
*1 The ground wiring of the converter should be connected to either the protective ground terminal in the equipment or
the ground terminal of the converter case.
Ground to earth, ground resistance: 100 Ω or less.
Figure 5.2
5.1.3
Wiring connection to the converter
Mounting of Cable Gland
For each cable connection opening of the converter, mount a conduit that matches the thread
size, or a cable gland.
EXA ZR402G
8-G1/2, 8-1/2NPT
or the like
(Wiring connection)
Ground terminal (M4)
Adapter for 1/2NPT thread
Figure 5.3
F5-4E.ai
Cable gland mounting
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<5. Wiring>
5-5
5.2 Wiring
5.2.1
Connection to Converter
To connect the wiring to the converter, proceed as follows:
(1) M4 screws are used for the terminals of the converter. Each cable should be terminated in
the corresponding size crimp-on terminals.
(2) When a rubber insulated glass braided wire is used for wiring to the detector, use a terminal
box. For wiring between the terminal box and the converter, basically use a cable that
withstand temperatures of at least 80°C.
NOTE
The above is to prevent moisture or corrosive gas from entering the converter.
Where the ambient environment of the detector and the converter is well-maintained, it is
permissible to connect the wiring from the detector directly to the converter with protection by
conduits.
WARNING
This wiring is to carry power for the heater. Be careful to wire the correct terminals, and be
careful not to ground or short circuit terminals when wiring, as otherwise the instrument may be
damaged.
5.2.2
Connection to Detector
When connecting the cable to the detector, proceed as follows:
(1) Mount cable glands or conduits of the specified thread size to the wiring connections of the
detector.
The detector may need to be removed in future for maintenance, so be sure to allow
sufficient cable length.
(2) If the ambient temperature at the location of wire installation is 75 to 150°C, be sure to use a
flexible metallic conduit for the wire. If a non-shielded “ 600 V silicon rubber insulated glass
braided wire” is used, keep the wire away from noise sources to avoid noise interference.
(3) The size of the terminal screw threads is M3.5. Each cable should be terminated in the
corresponding size crimp-on terminals contact (*1) respectively.
*1 If the ambient temperature at the detector installation site exceeds 60°C, use a “ bare crimp-on terminal”.
l Notice when closing the cover of the detector
NOTE
• Before opening the detector cover, loosen the lock screw. If the screw is not loosened first,
the screw will damage the cover, and the terminal box will require replacement.
When opening and closing the cover, remove any sand particles or dust to avoid gouging
the thread.
• After screwing the cover in the detector body, secure it with the lock screw.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<5. Wiring>
5-6
Lock screw
Detector cover
F5-8E.ai
Figure 5.4
5.2.3
Power and Grounding Wiring
This wiring supplies power to the converter and grounds the converter/detector.
ZR402G
Converter
Grounding to the ground terminal
on the converter case
ZR22G
Detector
L N G
Converter case
Jumper plate
FG terminal
Figure 5.5
Lock washer
Crimp-on terminal of
the ground wire
100 - 240 V AC
50/60 Hz
Ground
F5-10E.ai
Power and Grounding wiring
Power Wiring
Connect the power wiring to the L and N terminals of the converter. Proceed as follows:
(1) Use a 2-core or a 3-core cable.
(2) The size of converter terminal screw threads is M4. Each cable should be terminated
corresponding to crimp-on terminals.
Grounding Wiring
The ground wiring of the detector should be connected to the ground terminal of the detector
case. The ground wiring of the converter should be connected to either the ground terminal of the
converter case or the protective ground terminal in the equipment.
The grounding terminals of the detector and the converter are of size M4. Proceed as follows:
(1) Keep ground resistance to 100 Ω or less (equivalent JIS D style (Class 3)).
(2) When the ambient temperature of the wiring installation is 75 to 150°C for the wiring of the
detector, use wiring material with sufficient heat resistance.
(3) When connecting the ground wiring to the ground terminal of the converter case, be sure
that the lock washer is in contact with the case surface (see Figure 5.5.).
(4) Ensure that the jumper plate is connected between the G terminal and the protective ground
terminal of the converter.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
5.2.4
<5. Wiring>
5-7
Wiring for Power to Detector Heater
This wiring provides electric power from the converter to the heater for heating the sensor in the
detector.
(1) Ambient temperature of the detector: 75°C or less
ZR402G
Converter
ZR22G
Detector
HTR 7
HTR 8
HEATER
(2) Ambient temperature of the detector: exceeding 75°C
ZR22G
Detector
Terminal box
HTR 7
HTR 8
ZR402G
Converter
HEATER
Heat-resistant wiring
F5-7E.ai
Figure 5.6
Wiring for power to detector heater
Cable Specifications
Basically, cables (2 cores) that withstand temperatures of at least 80°C are used for this wiring.
When the ambient temperature of the detector exceeds 75°C, install a terminal box, and connect
to the detector using six-piece 600 V silicon rubber insulated glass braided wires.
5.2.5
Wiring for Detector Output
This wiring enables the converter to receive cell output from the detector, output from a
thermocouple and a reference junction compensation signal. Install wires that allow for 10 Ω of
loop resistance or less. Keep detector wiring away from power wiring.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<5. Wiring>
5-8
(1) Ambient temperature of the detector: 75°C or less
ZR402G
Converter
ZR22G
Detector
CELL(+)
CELL(+)
CELL(-)
CELL(-)
TC(+)
TC(+)
TC(-)
CJ(+)
TC(-)
CJ(-)
CJ(-)
CJ(+)
FG
Shielded cables
(2) Ambient temperature of the detector: exceeding 75°C
ZR22G
Detector
ZR402G
Converter
Terminal box
CELL(+)
CELL(+)
CELL(-)
CELL(-)
TC(+)
TC(+)
TC(-)
TC(-)
CJ(+)
CJ(+)
CJ(-)
CJ(-)
Heat-resistant wiring
Figure 5.7
FG
Shielded cables
F5-5E.ai
Wiring for detector output
CAUTION
If shielded cables cannot be used between the detector and the terminal box, for example,
when heat-resistant wiring is used, locate the detector and the terminal box as close together as
possible.
Cable Specifications
Basically, a cable (6-core) that withstand temperatures of at least 80°C is used for this wiring.
When the ambient temperature of the detector exceeds 75°C, install a terminal box, and connect
with the detector using six-piece 600 V silicon rubber insulated glass braided wires.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
5.2.6
<5. Wiring>
5-9
Wiring for Analog Output
This wiring is for transmitting 4 to 20 mA DC output signals to a device, e.g. recorder.
Maintain the load resistance including the wiring resistance at 550 Ω or less.
ZR402G
Converter
Receiver 1
1
2
AO-1(+)
AO-1(-)
AO-2(+)
AO-2(-)
Receiver 2
Shielded cable
FG
1
2
F5-9E.ai
Figure 5.8
Wiring for analog output
Cable Specifications
For this wiring, use a 2-core or a 4-core shielded cable.
Wiring Procedure
(1) M4 screws are used for the terminals of the converter. Each wire in the cable should be
terminated corresponding to crimp-on terminals. Ensure that the cable shield is connected
to the FG terminal of the converter.
(2) Be sure to connect “+” and “-” polarities correctly.
5.2.7
Contact Output Wiring
Contact outputs 1 to 3 can be freely assigned to “low limit alarm”, “high limit alarm”, etc. user
selectable, but the assignment of contact output 4 is fixed (“error output”). And the action (contact
closed on error output) also cannot be changed.
When using these contact outputs, install the wiring as follows:
ZR402G
Converter
Terminal box
Annunciator or the like
DO-1
DO-1
#1 Output
DO-2
DO-2
#2 Output
DO-3
DO-3
#3 Output
DO-4
DO-4
#4 Output
F5-11E.ai
Figure 5.9
Contact output wiring
Cable Specifications
Number of wire in cable varies depending on the number of contact used.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<5. Wiring>
5-10
Wiring Procedure
(1) M4 screws are used for the terminals of the converter. Each cable should be terminated
corresponding to crimp-on terminals.
(2) The capacities of the contact output relay are 30 V DC 3 A, 250 V AC 3 A. Connect a load
(e.g. pilot lamp and annunciator) within these limits.
5.2.8
Contact Input Wiring
The converter can execute specified function when receiving contact signals.
To use these contact signals, wire as follows:
ZR402G
Converter
Terminal box
DI-1
Contact input 1
DI-2
DI-C
Contact input 2
F5-14E.ai
Figure 5.10
Contact Input Wiring
Cable Specifications
Use 2-core or 3-core cable for this wiring. Depending on the number of input(s), determine which
cable to use.
Wiring Procedure
(1) M4 screws are used for the terminals of the converter. Each cable should be terminated
corresponding to crimp-on terminals.
(2) The ON/OFF level of this contact input is identified by the resistance. Connect a contact
input that satisfies the specifications in Table 5.2.
Table 5.2
Resistance
Identification of Contact Input ON/OFF
Closed
Open
200 Ω or less
100 kΩ or more
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<5. Wiring>
5.2.9
Wiring for ZR40H Automatic Calibration Unit
5-11
This wiring is for operating the solenoid valve for the zero gas and the span gas in the ZR40H
Automatic Calibration Unit, in a system where the calibration gas flow rate is automatically
controlled (e.g. System configuration 3). When installing this wiring, proceed as follows:
Wiring inlet
2-G1/2, Pg13.5, M20 or 1/2NPT
Span gas solenoid valve
Zero gas solenoid valve
ZERO
COMMON
SPAN
COMMON SPAN
ZERO
to ZR402G terminal "AC-COM"
to ZR402G terminal "AC-SPAN"
to ZR402G terminal "AC-ZERO"
Terminal screw : M4
F5-12E.ai
Figure 5.11
Automatic Calibration Unit
Cable Specifications
Use a 3-core cable for the above wiring.
Wiring Procedure
M4 screws are used for the terminals of the converter. Each cable should be terminated
corresponding to crimp-on terminals. M4 screws are used for the terminals of the solenoid valve
as well.
ZR402G
Converter
AC-Z
ZR40H Automatic
Calibration Unit
Zero
AC-S
AC-C
Span
F5-13.ai
Figure 5.12
Wiring for Automatic Calibration Unit
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
5.2.10
<5. Wiring>
5-12
Temperature Input Wiring
(Only for Humidity Analyzer)
When inputting the measurement gas temperature from external of the equipment, connect a
two-wire temperature transmitter. The relative humidity and dew point are acquired based on the
temperature signal from the connected transmitter, in the case where the setting is “Temperature
input selected” and “external input”. As for the wiring of the temperature transmitter and
thermocouples, refer to appropriate temperature transmitter instruction manual.
Converter
Temperature transmitter
AI+
+
AI-
-
FG
Shielded cables
F5-14E.ai
Figure 5.13
l Applicable Temperature Transmitter
Apply a temperature transmitter that is suit for the following interfaces:
Output signal: 4 to 20 mA DC, two-wire system
Maximum supply voltage from the analyzer:
24 V DC
Input resistance of the analyzer: Maximum 250 V (The load resistance of the transmitter is the
total of wiring resistance and input resistance.)
Temperature Transmitter Burnout
When outputting a burnout signal of the temperature transmitter with a contact output of the
analyzer, use “high-limit temperature alarm”. (Refer to Section 8.4, “Contact Output Setting.”) In
this case, set the burnout signal of the temperature transmitter to exceed the high limit (20 mA or
more).
Cable Specifications
Use a two-core shielded cable for wiring.
Wiring Procedure
(1) M4 screws are used for the converter terminals. Cables should be equipped with
appropriate crimp contacts. Ensure that the cable shield be connected to the FG terminal of
the converter.
(2) Be sure to connect “+” and “-” polarities correctly.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
6-1
<6. Components>
6.
Components
In this Chapter, the names and functions of components are described for the major equipment of
the EXAxt ZR Separate type Zirconia Oxygen/Humidity Analyzer.
In the figure listed in this manual, the example of the oxygen analyzer is shown mainly.
In the case of the humidity analyzer, unit indication may be different. Please read it appropriately.
6.1
ZR22G Detector
6.1.1
General-purpose Detector (except for ZR22G-015)
Terminal box,
Non explosion-proof JIS C0920 equivalent to IP44D.
Equivalent to NEMA 4X/IP66 (Achieved when the cable entry
is completely sealed with a cable gland in the recirculation
pressure compensated version.)
Flange
Used to mount the detector.
Select from JIS, ANSI, JPI or DIN
standard models.
Probe
This part is inserted in the furnace.
Select length from 0.4, 0.7, 1.0,
1.5, 2.0, 2.5, 3.0, 3.6, 4.2, 4.8 or 5.4 m.
Dust filter mounting screw
Calibration gas
pipe opening
Metal O-ring
Pipe support
Contact
U-shaped pipe
Bolt
Probe
Sensor (cell)
Filter
F6-1E.ai
Washer
Figure 6.1
General-purpose Detector
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
6.1.2
<6. Components>
6-2
High Temperature Detector (ZR22G-015)
Sample gas outlet
When a sample gas pressure is
negative, connect the ejector assembly.
When the sample gas is high
temperature and high pressure, and does
not fall below 700°C, connect a pressure
control valve (e.g. a needle valve).
(Refer to Section 3.2.1.)
Separate type
High Temperature Detector (ZR22G-015)
When the temperature of the sample gas is
between 700°C and 1400°C, mount this
detector with a ZO21P-H probe adapter.
Flange
Selectable from JIS, ANSI,
JPI or DIN
High Temperature Probe Adapter (ZO21P-H)
The probe is made of either SUS310S
or silicon carbide (SiC).
Its length is 0.5, 0.6, 0.7, 0.8, 0.9, 1.0 or 1.5m.
When using an SiC probe, mount it vertically
downward.
F6.2E.ai
Figure 6.2
High Temperature Detector
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
6.2
6-3
<6. Components>
ZR402G Converter
Complete Operation Display
■ Interactive operations along with operation display.
■ A variety of display modes - enabling you to select
the operation mode freely.
■ Back-lit LCD allows viewing even in the darkness.
■ Error codes and details of errors can be checked in
the field without the need to refer to appropriate
instruction manual.
Typical Converter Displays
● Example of basic display
This display enables you to operate the analyzer
while checking data on the display.
● Example of trend display
— displays data changes
Self-testing suggests countermeasures
for problems
If a problem occurs, the liquid-crystal display will
provide an error code and description of the
problem. This enables prompt and appropriate
corrective action to be taken.
During calibration, you can check the stabilized,
display data while viewing oxygen trend data,thus
providing highly reliable calibration.
Error code
Reason for error
E--1
Cell failure
E--2
Abnormal heater temperature
E--3
Defective A/D converter
E--4
Faulty EEPROM
ALARM1
Abnormal oxygen concentration
ALARM2
Abnormal moisture
ALARM3
Abnormal mixing ratio
ALARM6
Abnormal zero calibration ratio
ALARM7
Abnormal span calibration ratio
ALARM8
Stabilization time over
● Example of setting data display
— displays data changes
Maintenance
Display setup
Calibration setup
Blow back setup
mA-output loop check
Contact check
Enter
● One-touch interactive display operation
● User-friendly design providing easy operation
F7.6E.ai
without having to use the instruction manual
Figure 6.3
F6-3E.ai
Converter
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
6.3
6-4
<6. Components>
Touchpanel Switch Operations
6.3.1
Basic Panel and Switch
The converter uses a touchpanel switch which can be operated by just touching the panel
display. Figure 6.4 shows the Basic panel display. The switches that appear in the switch display
area vary depending on the panel display, allowing all switch operations. Table 6.1 shows the
switch functions.
Tag name display area
Tag:
Primary value
Secondary value
Tertiary value
21.0
Switch display area
% O2
Alarm and error
display area
17.43mA -Output1
17.43mA -Output2
F7.3E.ai
Figure 6.4
Basic Panel Display
Tag name display area: Displays the set tag name (Refer to Section 10.3.2, Entering Tag Name).
Primary to tertiary value display items: Displays the selected item. (Refer to Section 7.9, Setting
Display Item.)
Switch display area: Displays switches and functions selected according to the panel display.
Alarm and error display area: Displays an error if an alarm or error occurs. If you touch this
area, the details of the error or alarm are then displayed.
Table 6.1
Switches and Their Functions
Home key:
Returns to the Execution/Setup display.
Enter
Enter key:
Enters the input value and sets up the
selected item.
Reject key:
Moves back to the previous display.
Setup key:
Used to enter the Execution/Setup display.
Cursor key:
Moves the cursor down.
Detailed-data key:
Displays the analog input value.
Graph display key:
Displays a trend graph.
Cursor:
Points the cursor at the currently
selected item.
Alarm:
Displayed if an alarm arises.
Error:
Displayed if an error occurs.
T7-1E.ai
6.3.2
Display Configuration
Figure 6.5 shows the configuration. A password the displays positioned below enables display
“Execution/Setup” to be protected. If a password has not been set, press the [Enter] key to
proceed to the next panel display. The [Home] key enables you to return to “Execution/Setup
display” from any panel display.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
6-5
<6. Components>
[Basic panel display]
[Detailed-data display]
[Execution/Setup display]
[Trend Graph display]
[Password Entry display]
Calibration execution
Blow back execution
Returns to the
Execution/Setup
display.
Returns to the
Execution/Setup
display.
Reset panel display
Returns to the
Execution/Setup
display.
Maintenance panel
display
Display setup
Display item
Trend graph
Auto return time
Language
Calibration setup
Mode
Points
Zero and Span gas
conc.
Timing
Blow back setup
mA-output loop check
Contact check
Returns to the
Execution/Setup
display.
Commissioning
(Setup) display
mA-output setup
mA-output1
mA-output2
Set presets
Alarms setup
Hysteresis
Contact delay
Setpoints
Contact Output setup
Contact Input setup
Others
Clock
Averaging
Fuel setup
Passwords
Defaults
Returns to the
Execution/Setup
display.
F7.3.1E.ai
Figure 6.5
6.3.3
Display Configuration
Display Functions
Individual panel displays in the display configuration provide the following functions:
(1) Basic panel display: Displays the values measured in three selected items (see Section 7.9,
Setting Display Item).
(2) Execution/Setup display: Selects the calibration, maintenance and setup items.
(3) Detailed-data display: This allows you to view such detailed data as the cell electromotive
force and cell temperature (see Section 10.1, Detailed-data Display, later in this manual).
(4) Trend Graph display: Displays a trend graph (see Section 10.2, Trend Graph, later in this
manual).
(5) Calibration execution: Makes zero and span calibrations (see Chapter 9, Calibration, and
the associated sections later in this manual).
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<6. Components>
6-6
(6) Blow back execution: Executes a “Blow back setup” (see Section 10.4, Blow Back, later in
this manual).
(7) Reset panel display: If an error arises, you can restart the equipment from this display (for
more details, see Section 10.6, Reset, later in this manual).
(8) Maintenance panel display: Sets the data for equipment maintenance or makes a loop
check.
(9) Commissioning (Setup) display: Sets up the operation data. (For details, see Chapter 8,
Detailed Data Settings, and the associated sections later in this manual.)
6.3.4
Entering Numeric and Text Data
This section sets out how to enter numeric and text data. If only numeric values are entered,
a numeric-data entry display as in Figure 6.6 then appears. Press the numeral keys to enter
numeric values. If those values include a decimal point as in Figure 6.6, the decimal point need
not be entered because the decimal point position is already fixed, so just enter 00098.
New value:
0
000.98 % O2
7
8
9
4
5
6
1
2
3
Enter
F7.4E.ai
Figure 6.6
Numeric-data Entry Display
To enter a password (in combination with text data, numeric values and codes), the alphabetic
character entry panel display first appears. If you press any numeral key [0 - 9], the current
display then changes to the numeric-value entry panel display, enabling you to enter numeric
values. If you press the [other] key, the current display then changes to the code-entry display,
enabling you to enter codes. These displays alternate between the three. Figure 6.7 shows the
relationship between these three displays. Three alphabetic characters and three codes are
assigned for each individual switch. If the alphabetic character key is pressed and held, three
characters appear in turn. Move the cursor to the desired character and release the key to enter
it. If an incorrect character is entered, move the cursor to re-enter the characters. The following
shows an example of entering “abc%123.”
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<6. Components>
Operation
Press the [ABC] key once.
Press and hold the [ABC] key.
6-7
Display
A_
AA
B
C
Release the [ABC] key when the character B
appears in the cursor position.
Enter the character C in the same manner
as above.
Press the [other] key.
AB_
Press and hold the [$%&] key and enter “%.”
Then press the [0-9] key.
ABC%_
Enter the numeric characters 1, 2 and 3 in turn.
Press the [Enter] key to complete the entry.
ABC%123_
ABC_
Siki7.4E
Enter your password
0-9
A
B C
D
E F
G
H I
other
J
K L
M
N O
P
Q R
Space
- $
S
T U
V
W X
Y
Z @
Enter your password
Enter
Enter your password
A-Z
7
8
9
A-Z
!
0 #
$
% &
'
( )
other
4
5
6
0-9
*
+ ,
–
. /
:
; <
0
1
2
3
>
? @
[
\ ]
^
_ `
Enter
[
{
}
Enter
F7.5E.ai
Figure 6.7
Text Entry Display
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
6.4
<6. Components>
6-8
ZA8F Flow Setting Unit, ZR40H Automatic
Calibration Unit
Reference gas flow
setting valve
Span gas flow
setting valve
Zero gas flow
setting valve
Flowmeter for
reference gas
Flowmeter for
calibration gas
Figure 6.8
F6-4E.ai
ZA8F Flow Setting Unit
Flowmeter for
Calibration gas
Flowmeter for
Reference gas
Reference gas flow
setting valve
Terminal Box
Made in Japan
REF.OUT CAL.OUT SPAN IN ZERO IN
Calibration gas flow
setting valve
F6.5.ai
Figure 6.9
ZR40H Automatic Calibration Unit
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<7. Startup>
7-1
7. Startup
The following describes the minimum operating requirements — from supplying power to the
converter to analog output confirmation to manual calibration.
System tuning by the HART communicator, refer to IM11M12A01-51E “ HART Communication
Protocol ”.
In the figure listed in this manual, the example of the oxygen analyzer is shown mainly.
In the case of the humidity analyzer, unit indication may be different. Please read it appropriately.
7.1
Startup Procedure
The startup procedure is as follows:
CAUTION
If you connect Model ZO21D* detectors, then you need to change the detector parameters.
Before connecting power, refer to: Section 7.6 Confirmation of Detector Type Setting.
Check settings
Check piping & wiring
Check valve type setting
If connecting ZO21D
Briefly power on
Set detector to ZO21D
(front panel)
If detector is ZR22
Power OFF
Permanent power wiring
Start set parameters
Warmup, then calibrate
Figure 7.1
7.2
F07-1E.ai
Startup Procedure
Checking Piping and Wiring Connections
Check that the piping and wiring connections have been properly completed in accordance with
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<7. Startup>
7-2
Chapter 4, “Piping,” and Chapter 5, “Wiring.”
7.3
Checking Valve Setup
Set up valves and associated components used in the analyzer system as follows:
(1) If a stop valve is used in the detector’s calibration gas inlet, fully close this valve.
(2) If instrument air is used as the reference gas, adjust the air-set secondary pressure so that
an air pressure equals sample gas pressure plus approx. 50 kPa (or sample gas pressure
plus approx. 150 kPa when a check valve is used, maximum pressure rating is 300 kPa) is
obtained. Turn the reference gas flow setting valve in the flow setting unit to obtain a flow
of 800 to 1000 ml/min. (Turning the valve shaft counterclockwise increases the rate of flow.
Before turning the valve shaft, if the valve has a lock nut, first loosen the lock nut.) After
completing the valve setup, be sure to tighten the lock nut.
NOTE
The calibration gas flow setting is described later. Fully close the needle valve in the flow setting
unit.
7.4
Supplying Power to the Converter
CAUTION
To avoid temperature changes around the sensor, it is recommended that (rather than turning it
on and off) power be continuously supplied to the Oxygen Analyzer if it is used in an application
where it is used periodically.
It is also recommended to flow a span gas (instrument air) beforehand.
Supply power to the converter. A display as in Figure 7.2, which indicates the detector’s sensor
temperature, then appears. As the heat in the sensor increases, the temperature gradually rises
to 750°C. This takes about 20 minutes after the power is turned on, depending somewhat on the
ambient temperature and the sample gas temperature.
After the sensor temperature has stabilized at 750°C, the converter is in measurement mode.
The display panel then displays the oxygen concentration as in Figure 7.3. This is called the
Basic panel display.
Tag:
Tag:
319
°C
% O2
17.43mA -Output1
17.43mA -Output2
4.00mA -Output1
4.00mA -Output2
Hold
Warmup
Figure 7.2
Display During Warm-up
21.0
Figure 7.3
Measurement Mode Display
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
7.5
7-3
<7. Startup>
Confirmation of Converter Type Setting
This converter can be used for both the Oxygen Analyzer and the Humidity Analyzer.
Before setting the operating data, be sure to check that the desired converter model has been
set.
Note that if the converter type setting is changed, the operating data that have been set are then
initialized and the default settings remain. To set the desired operating data, follow these steps:
(1) Press the [Setup] key.
(2) Use the [▼] key to select “Commissioning” and press the [Enter] key.
(3) In the password display, enter the [Enter] key. If the password is to be set again, enter the
new password (for details, see Section 8.6.6, Setting Passwords, later in this manual).
(4) The Commissioning display shown in Figure 7.4 appears. Select “Basic setup” and press
the [Enter] key.
(5) The Basic setup display shown in Figure 7.5 then appears. Confirm the currently set
converter type. If the Humidity Analyzer option /HS was selected at the time of purchase, the
converter was set for high temperature humidity use before shipment.
(6) If the converter type is to be changed, press the [Enter] key. The display shown in Figure 7.6
then appears.
(7) Use the [▼] key to select the type of equipment. Then press the [Enter] key to complete the
converter selection.
(8) If the type of converter is changed after setting the operating data, those data are then
initialized and the default settings remain. Reset the operating data to meet the new type of
equipment.
Commissioning
Basic setup
Basic setup
Contact setup
This instrument is for
Oxygen
Please note:changing this
Basic setting will return ALL
other setting to Default!!
Others
Detector is ZR22
mA-output setup
Alarms setup
Select measure gas: Wet
Enter
F7.6E.ai
Figure 7.4
Commissioning Display
Enter
F7.7E.ai
Figure 7.5
Basic Setup
Basic setup
This instrument is for
Oxygen
Humidity
Please note:changing
this
Basic setting will return ALL
other setting to Default!!
Detector is ZR22
Select measure gas: Wet
Enter
F7.8E.ai
Figure 7.6
Equipment Setup
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
7.6
7-4
<7. Startup>
Confirmation of Detector Type Setting
Check that the detector in Figure 7.5 is the one for this equipment.
CAUTION
• If this converter is to be used in conjunction with the ZO21D, the power requirements are
limited to 125 V AC or less, 50 or 60 Hz (it cannot be used with a 125 V or greater, or in the
EEC).
• If detector settings are to be changed, first disconnect the wiring connections between the
detector and the converter. Then change detector settings appropriately.
7.7
7.8
Selection of Sample Gas
Combustion gases contain moisture created by burning hydrogen in the fuel. If this moisture is
removed, the oxygen concentration might be higher than before. You can select whether the
oxygen concentration in a wet gas is to be measured directly, or compensated for its dry-gas
value before use. Select the “Select measure gas: Wet” in Figure 7.6 to select either wet or dry
gas.
Output Range Setting
This section sets forth analog output range settings. For details, consult Section 8.1, Current
Output Setting, later in this manual.
7.8.1
Oxygen Analyzer - Minimum Current (4 mA) and
Maximum Current (20 mA) Settings
To set the minimum and maximum current settings, follow these steps:
(1) Select the “Commissioning” from the Execution/Setup display.
(2) From the Commissioning (Setup) display, select “mA-output setup”; the display shown in
Figure 7.7 then appears.
(3) Select “mA-output1” from the mA-outputs display. The “mA-output1 range” display shown in
Figure 7.8 then appears.
(4) In the display shown in Figure 7.8, select “Min. oxygen conc” and press the [Enter] key to
display the numeric-data entry display. Enter the oxygen concentration at a 4-mA output;
enter [010] for a ten-percent oxygen concentration measurement.
(5) Also in Figure 7.8, select “Max. oxygen conc” at a 20-mA output. Enter the appropriate
maximum oxygen concentration (at the 20-mA output) in the same manner as in step 4
above.
(6) Set “mA-output2” in the same manner as in the appropriate steps above.
mA-output1 range
mA-outputs
Both outputs are 4-20mA
mA-output1
mA-output2
Set presets
Enter
F7.9E.ai
Figure 7.7
Setting “mA-outputs”
Parameter: Oxygen
Set range
Min. oxygen conc:
0
%O2
Max. oxygen conc:
25
%O2
Output damping:
0s
Output mode: Linear
Enter
F7.10E.ai
Figure 7.8
Setting “mA-output1 range”
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
7.8.2
7-5
<7. Startup>
Humidity Analyzer - Analog Output Setting
Select any one of the analog output settings — Oxygen, Humidity, and Mixing from the mAoutput range display. If the /HS option is specified at the time of purchase, the equipment is a
humidity analyzer. For other than this setting, the analyzer is an oxygen analyzer. If a mixed
measurement is required, change the current output setting as given below. If the humidity
analyzer is specified in the above detector type setting, the analog output is set to “humidity”
when the data initialization is attempted.
(1) Select the Setup from the Execution/Setup display.
(2) From the Commissioning (Setup) display, select “mA-output setup”; the display shown in
Figure 7.9 then appears.
(3) Select “mA-output1” from the “mA-outputs” display. The “mA-output1 range” display shown
in Figure 7.10 then appears.
(4) Select the Parameter, then the Oxygen, Humidity, and Mixing appear in the mAoutput1
range display (Figure 7.11). Select “Mixing” and press the [Enter] key.
(5) The “Mixing” in Figure 7.12 appears.
(6) Perform the same setting in mA-outpout2 if necessary.
mA-outputs
mA-output1 range
Both outputs are 4-20mA
mA-output1
mA-output2
Set presets
Enter
F7-7E.ai
Figure 7.9
mA-outputs Setting Display
mA-output1 Range Display
mA-output1 range
Parameter: Oxygen
Humidity
Humidity
Set range
Mixing
Min. humidity
conc:
0
%H2O
Max. humidity conc:
25
%H2O
Output damping:
0s
Output mode: Linear
Parameter: Mixing
Set range
Min. mixing ratio:
0. 000 kg/kg
Max. mixing ratio:
0. 200 kg/kg
Output damping:
0s
Output mode: Linear
Enter
Figure 7.11
mA-output1 Range Display
Enter
F7-8E.ai
Figure 7.10
mA-output1 range
F7-9E.ai
7.8.3
Parameter: Humidity
Set range
Min. humidity conc:
0
%H2O
Max. humidity conc:
25
%H2O
Output damping:
0s
Output mode: Linear
Enter
F7-10E.ai
Figure 7.12
Mixing Ratio Display
Humidity Analyzer - Minimum Current (4 mA) and
Maximum Current (20 mA) Settings
To set the analog output range, follow these steps:
(1) Select the Setup from the Execution/Setup display.
(2) From the Commissioning (Setup) display, select the “mA-output setup.” Select “mAoutput1”
from the “mA-outputs” display. The “mA-output1 range” that displays “Parameter: Humidity,”
as shown in Figure 7.10 then appears.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<7. Startup>
7-6
(3) To set the minimum humidity at 4 mA, choose the Min. humidity conc. To set the maximum
humidity at 20 mA, choose the Max. humidity conc.
(4)
To set 50% H2O, type in 050 and press the [Enter] key.
(5)
Set “mA-output2” in the same manner as in the appropriate steps above.
NOTE
Analog output range settings should be limited. For more details, consult Section 8.1, “Current
Output setting,” later in this manual.
7.9
Setting Display Item
7.9.1
Oxygen Analyzer - Setting Display Item
This section briefly describes the display item settings shown in Figure 7.13, “Basic Panel
Display.”
Tag:
Primary value
Secondary value
Tertiary value
21.0
%O2
17.43mA -Output1
17.43mA -Output2
F7.11E.ai
Figure 7.13
Basic Panel Display
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
Then select “Maintenance” in the Execution/Setup display.
(2) Select the “Display setup” from the Maintenance panel display (Figure 7.14).
The Display setup display (Figure 7.15) then appears.
(3) In the above Display setup display, select the “Display item”. The Display item display
(Figure 7.16) then appears. From this display, select the “Primary value” and press the
[Enter] key to display the “Display item” selection display (Figure 7.17).
(4) Select the Secondary and Tertiary values in the same manner as in the steps above.
(5) Consult Table 7.1, Display Items, enabling the selection of display items in individual display
areas.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
7-7
<7. Startup>
Maintenance
Display setup
Display item
Display setup
Calibration setup
Blow back setup
Trend graph
Auto return time:
mA-output loop check
Contact check
Language:
0min
English
Enter
Enter
F7.12.ai
Figure 7.14
Maintenance Panel Display
F7.13.ai
Figure 7.15
Display item
Display item
Primary value:
Oxygen
Secondary value:
mA-output1
Tertiary value:
mA-output2
Tag name:
Enter
F7.14.ai
Figure 7.16
Display Item Display
Table 7.1
Display Items
Item
Primary value
Oxygen
concentration
○
Air ratio
Moisture quantity
Output 1 item
○
Output 2 item
○
Current output 1
Current output 2
*1:
Display Setup
Primary value:
Oxygen
Oxygen
Item of output1 damping
Secondary
value:
Item of output2
damping
mA-output1
Tertiary value:
mA-output2
Tag name:
Enter
F7.15.ai
Figure 7.17
Secondary and
tertiary values
Display Item Selection
Display
○
○
○
○
Oxygen concentration during measurement
○
○
○
Oxygen concentration with the equipment set for oxygen
analyzer (See *1 below.)
Current computed air ratio
Moisture quantity (%H2O) in the exhaust gas
Oxygen concentration with the equipment set for oxygen
analyzer (See *1 below.)
Current value output from analog output 1
Current value output from analog output
If an analog output damping constant is set, the oxygen concentration display then includes these settings.
l About the Air ratio:
“Air ratio” is defined as the ratio of (the amount of air theoretically required to completely burn all
the fuel) to (the amount of air actually supplied).
For this equipment, the air ratio will be obtained in a simplified way by measuring the oxygen
concentration in the exhaust gas. The air ratio may be expressed mathematically by:
m = { 1 / ( 21- Oxygen concentration ) } 21
If you use the air ratio data for estimating the combustion efficiency, etc., check that no air is
leaking in beforehand and that the measured value has not been affected by any interference
gas (CH4, CO, H2, etc.).
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<7. Startup>
7-8
l About moisture quantity:
The moisture quantity in the exhaust gas is calculated based on the parameters of the fuel
setting (refer to Section 8.6.3, Setting Fuels, later in this manual). The moisture content may be
expressed mathematically by:
Moisture quantity = {(water vapor content per fuel unit quantity) + (water content in air)}/ total
amount of exhaust gas
= { Gw + ( 1.61 Z Ao m )} / { X + ( Ao m )}
where,
Gw = water vapor content in exhaust gas, m3/kg (m3/m3)
Z = Ambient absolute humidity, kg/kg
Ao = Ideal air amount, m3/kg (m3/m3)
m = Air ratio
X = Fuel coefficient
For details on each parameter, refer to Section 8.6.3, Setting Fuels.
7.9.2
Humidity Analyzer - Setting Display Item
This section briefly describes the display item settings shown in Figure 7.18, “Basic Panel
Display.” If the humidity analyzer /HS option was specified at the time of purchase, the primary
value has set “Humidity.” If a mixing ratio is to be measured, change the current primary value
following the steps below.
Additionally, if the “humidity” has been selected in the Detector Type Setting in Section 7.6, earlier
in this manual, the primary value is set to the humidity and the secondary and tertiary values are
current outputs 1 and 2, respectively when data initialization is performed.
Tag:
Primary value
Secondary value
Tertiary value
0.0
%H2O
4.00mA -Output1
4.00mA -Output2
F7-11E.ai
Figure 7.18
Basic Panel Display
(1) Press the Setup key in the basic panel display to display the Execution/Setup display.
Then select Maintenance in the Execution/Setup display.
(2) Select the Display setup from the Maintenance panel display (Figure 7.19). The Display
setup display (Figure 7.20) then appears.
(3) In the above Display setup display, select the Display item. The Display item display (Figure
7.21) then appears. From this display, select the Primary value and press the [Enter] key to
display the Display item selection display (Figure 7.22).
(4) Select the Secondary and Tertiary values in the same manner as in the steps above.
(5) Consult Table 7.2, “Display Items,” enabling the selection of display items in individual
display areas.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
7-9
<7. Startup>
Maintenance
Display setup
Display item
Display setup
Calibration setup
Blow back setup
Trend graph
Auto return time:
mA-output loop check
Contact check
Language:
0min
English
Enter
Enter
F7-12.ai
Figure 7.19
Maintenance Panel Display
F7-13.ai
Figure 7.20
Display item
Display Setup
Display item
Primary value:
Humidity
Secondary value:
mA-output1
Tertiary value:
mA-output2
Tag name:
Primary value:
Oxygen
Oxygen
Humidity
Secondary
value:
Mixing
mA-output1
Item
of output1 damping
Item of value:
output2 damping
Tertiary
mA-output2
Tag name:
Enter
Enter
F7-14.ai
Figure 7.21
Display Item Display
Table 7.2
Display Items
Item
Oxygen
concentration
Humidity
Mixing ratio
Primary value
○
○
○
Relative humidity
Dew point
Output 1 item
○
Output 2 item
○
Current output 1
Current output 2
*1:
F7-15.ai
Figure 7.22
Secondary and
tertiary values
Display Item Selection
Display
○
○
○
○
○
○
Oxygen concentration during measurement
○
○
○
Oxygen concentration with the equipment set for oxygen
analyzer (See *1 below.)
Humidity (%H2O) in the exhaust gas
Mixing ratio during measurement
Relative humidity calculated from the measured value
Dew point calculated from the measured value
Oxygen concentration with the equipment set for oxygen
analyzer (See *1 below.)
Current value output from analog output 1
Current value output from analog output
If an analog output damping constant is set, the oxygen concentration display then includes these settings.
NOTE
For the relative humidity and dew-point calculations, appropriate operation parameters should
be entered. For details on the parameters, consult Section 8.6.3, “Setting Measurement Gas
Temperature and Pressure,” later in this manual.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
7.10
<7. Startup>
7-10
Checking Current Loop
The set current can be output as an analog output.
(1) Press the [Setup] key on the Basic panel display to display the Execution/Setup display.
Then select “Maintenance” in the Execution/Setup display.
(2) Select “mA-output loop check” in the Maintenance panel display to display the mA - output
loop check display, enabling you to check “mA-output1” and “mA-output2”.
Select the desired output terminal for current-loop checking (see Figure 7.23).
(3) At the time of entering the numeric-data entry display, the output current will change to 4 mA
(default value). If the desired current is entered, the corresponding output will be provided.
mA - output loop check
mA - output 1
mA - output 2
Enter
F7.15.1E.ai
Figure 7.23
“mA - output loop check” Panel Display
7.11
Checking Contact I/O
7.11.1
Checking Contact Output
Conduct the contact input and output checking as well as operational checking of the solenoid
valves for automatic calibration.
To check the contact output, follow these steps:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
Select “Maintenance” in that display.
(2) Select “Contact check” then “Output contacts” in the Maintenance panel display to display
the Output contacts display (see Figure 7.24).
(3) In this display, select the desired output contact for checking. The display, which enables
the closing and opening of contacts, then appears. Use the display to conduct continuity
checking.
Output contacts
Output contact 1 : Open
Output contact 2 : Open
Output contact 3 : Open
Output contact 4 : Open
Enter
F7.15.2E.ai
Figure 7.24
Output Contact Check Panel Display
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<7. Startup>
7-11
NOTE
If you conduct an open-close check for contact output 4, Error 1 or Error 2 will occur.
This is because the built-in heater power of the detector, which is connected to contact output 4,
is turned off during the above check. So, if the above error occurs, reset the equipment or turn the
power off and then back on to restart (refer to Section 10.6, Reset, later in this manual).
7.11.2
Checking Calibration Contact Output
The calibration contacts are used for solenoid valve drive signals for the ZR40H Automatic
Calibration Unit. When using the ZR40H Automatic Calibration Unit, use the calibration contact
output to check that the wiring connections have been properly completed and check equipment
operation.
(1) Referring to Section 7.11.1, display the contact check display.
(2) Select the “Calibration contacts” to display the panel display as Figure 7.25 shows.
(3) Open the “Zero gas contact” and the “Span gas contact”. This will help check the automatic
calibration unit and wiring connections.
Calibration contacts
Zero gas contact : Open
Span gas contact : Open
Enter
F7.15.3E.ai
Figure 7.25
Calibration Contact Check Display
NOTE
“Open” and “Closed” displayed on the Calibration contacts display indicate actions of drive
contacts and are opposite to the valve open and close actions. If “Open” is displayed on the
Calibration contacts display, no calibration gas flows. If “Closed” is displayed on that display,
calibration gas flows.
7.11.3
Checking Input Contacts
(1) Referring to Section 7.11.1, display the contact check display.
(2) Display the “Input contacts” check display as Figure 7.26 shows. The “Open” or “Closed”
input contact in the display shows the current contact input terminal status and the display
changes according to the contact status. Using this enables you to check that the wiring
connections have been properly completed.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<7. Startup>
7-12
Input contacts
Input contact 1 : Open
Input contact 2 : Open
Enter
F7.15.4E.ai
Figure 7.26
Input Contact Check Display
7.12 Calibration
To calibrate this instrument, the procedure is to measure zero gas and span gas and set the
instrument to read the known concentrations. The procedure for both zero and span calibration,
or for either zero or span calibration, can be performed manually from the touch display, or can
be performed semi-automatically using contact signal inputs to start calibration, (allowing preset
calibration and stabilization times), or it can be performed automatically at preset intervals.
There are three types of calibration procedures available:
(1) Manual calibration conducting zero and span calibrations, or either of these calibrations in
turn.
(2) Semi-automatic calibration which uses the touchpanel or a contact input signal and
conducts calibration operations based on a preset calibration time and stable time.
(3) Automatic calibration conducted at preset intervals.
Manual calibration needs the ZA8F Flow Setting Unit to allow manual supply of the calibration
gases. Semi-automatic and automatic calibrations need the ZR40H Automatic Calibration Unit
to allow automatic supply of the calibration gases. The following sections set forth the manual
calibration procedures. For details on semi-automatic and automatic calibrations, consult
Chapter 9, Calibration, later in this manual.
7.12.1
Calibration Setup
Mode Setting
For the mode setting, do the following:
Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
Select “Maintenance” in the Execution/Setup display to display the Maintenance panel display.
Then select “Calibration setup” to display the Calibration setup display as Figure 7.27 shows.
Select Mode in this panel, and then select “Manual”, “Semi_Auto” or “Auto”.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<7. Startup>
7-13
Calibration setup
Mode: Manual
Points: Both
Zero gas conc:
1.00%
Span gas conc: 21.00%
Timing
Enter
F7.16E.ai
Figure 7.27
Calibration Setup Display
Calibration Setting Procedures
Select “Points” (calibration procedure) in the Calibration setup display to display the “Both, Span,
Zero” selection display. In this display, select “Both.”
Calibration Gas Concentration Setting
(1) Zero gas concentration
If “Zero gas conc” is selected in the Calibration setup display, the numeric-data entry display
then appears.
Use this display to enter an oxygen concentration value for the zero gas calibration;
if the oxygen concentration is 0.98 vol%O2, enter 00098.
(2) Span gas concentration
With “Span gas conc” selected in the Calibration setup display, display the Numeric - data
Entry display and enter an oxygen concentration value for the span gas calibration; If
instrument air is used, enter 02100 for a 21 vol%O2 value. When using the ZO21S Standard
Gas Unit (for use of the atmospheric air as a span gas), use a portable oxygen analyzer to
measure the actual oxygen concentration, and then enter it.
NOTE
• If instrument air is used for the span gas, dehumidify the air by cooling to a dew point of - 20
ºC and remove any oil mist or dust.
• If dehumidification is insufficient, or polluted air is used, a measurement accuracy may be
adversely affected.
7.12.2
Manual Calibration
Preliminary
Before performing manual calibration, be sure that the ZA8F Flow Setting Unit zero gas flow
setting valve is fully closed. Open the zero gas cylinder pressure reducing valve so that the
secondary pressure equals sample gas plus approx. 50 kPa (or sample gas pressure plus
approx. 150 kPa when a check valve is used, maximum pressure rating is 300 kPa).
This applies even if you are using the ZR40H Automatic Calibration Unit.
Calibration Procedures
This manual assumes that the instrument air is the same as the reference gas used for the span
gas. Follow the steps below to conduct manual calibration:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
7-14
<7. Startup>
Then select “Calibration” in the Execution/Setup display.
Calibration
Manual calibration
Check the span gas
Manual calibration
concentration value
Next
Cancel calibration
Change value:
21.00%
Enter
F7.17E.ai
Figure 7.28
Calibration Display
Enter
F7.18E.ai
Figure 7.29
Manual Calibration
(2) Press the [Enter] key to select “Span calibration”. The Manual calibration display shown
in Figure 7.29 then appears. Check that the oxygen concentration for the span gas in this
display coincides with the oxygen concentration in the calibration gas actually used. If the
check results are assumed to be OK, select “Next” in the Manual calibration display.
(3) Follow the display message in Figure 7.30 to turn on span gas flow. Open the span gas flow
valve for the Flow Setting Unit by loosening the valve lock nut and slowly turning the valve
shaft counterclockwise to flow the span gas at 600 ± 60 ml/min. Use the calibration gas
flowmeter to check the flow.
Manual calibration
Manual calibration
Open span gas valve.
Set flow span gas to
600ml/min.
21.00%
0.5min./div
Valve opened
1.00%
Cancel calibration
21.0 %O2
Enter
F7.19E.ai
Figure 7.30
Span gas Flow Display
CAL. TIME
Enter
F7.20E.ai
Figure 7.31 Span gas Trend Graph
(for Manual Calibration)
(4) If “Valve opened” is selected as in Figure 7.30, an oxygen concentration trend graph (with
the oxygen concentration being measured) appears (see Figure 7.31). The CAL.TIME in
the bottom area of the panel flashes. Observe the trend graph and wait until the measured
value stabilizes in the vicinity of 21% on the graph. At this point, calibration has not yet been
executed yet, so even if the measured value is above or below 21%, no problem occurs.
(5) After the measured value has stabilized, press the [Enter] key to display the “Span
calibration complete” display shown in Figure 7.32. At this point, the measured value is
corrected to equal the span gas concentration setting. Close the span gas flow valve. The
valve lock nut should be tightened completely so that the span gas does not leak.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
7-15
<7. Startup>
Manual calibration
Manual calibration
Check the zero gas
Span calibration
Close the span gas valve.
concentration value
Next
Cancel calibration
Zero calibration
End
Change value:
1.00%
Enter
F7.21E.ai
Enter
F7.22E.ai
Figure 7.32 Span Calibration Complete
Figure 7.33
Zero Calibration
Start Display
Zero Gas Concentration
Check Display
(in Manual Calibration)
(6) Select “Zero calibration” as in Figure 7.32 to display the zero gas concentration check
display (Manual calibration) . Check that the zero gas oxygen concentration value and the
calibration gas oxygen concentration value agree. Then select “Next” as in Figure 7.33.
(7) Follow the instructions in the display as in Figure 7.34 to turn on the zero gas flow. To do
this, open the zero gas flow valve for the Flow Setting Unit and adjust that valve to obtain a
flow of 600 ± 60 ml/min. The valve should be adjusted by loosening its lock nut and slowly
turning the valve shaft counterclockwise. Use the calibration gas flowmeter to check the
flow.
Manual calibration
Manual calibration
Open zero gas valve.
Set flow zero gas to
600ml/min.
0.5min./div
21.00%
Valve opened
Cancel calibration
1.00%
21.0%O2
Enter
F7.23E.ai
Figure 7.34 Zero Gas Flow Display
(for Manual calibration)
CAL. TIME
Enter
F7.25E.ai
Figure 7.35
Zero Gas, Trend Graph
(8) If “Valve opened” is selected as in Figure 7.34 , an oxygen concentration trend graph (with
the oxygen concentration being measured) appears (see Figure 7.35). The CAL.TIME in
the bottom area of the panel flashes. Observe the trend graph and wait until the measured
value stabilizes in the vicinity of the zero gas concentration on the graph. At this point, no
calibration has been executed yet, so even if the measured value is above or below the zero
gas concentration value, no problem occurs.
(9) After the measured value has stabilized, press the [Enter] key to display the “Zero
calibration complete” display shown in Figure 7.36. At this point, the measured value is
corrected to equal the zero gas concentration setting. Close the zero gas flow valve. The
valve lock nut should be tightened completely so that the zero gas does not leak.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<7. Startup>
7-16
Manual calibration
Zero calibration
Close the zero gas valve.
Span calibration
End
Enter
F7.26E.ai
Figure 7.36
Zero Calibration Complete Display
(10) Select “End” in the display as shown in Figure 7.36. An oxygen concentration trend graph
(with the oxygen concentration being measured) appears and HOLD TIME then flashes.
This time is referred to as the output-stabilization time. If the HOLD TIME has been set
with the output hold setting, the analog output remains held (refer to Section 8.2, Output
Hold Setting, later in this manual). Manual calibration is completed when the preset hold
(output stabilization) time elapses. This hold (output stabilization) time is set to 10 minutes
at the factory before shipment. If you press the [Enter] or [Reject] key within the hold (output
stabilization) time, manual calibration is then completed.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<8. Detailed Data Setting>
8.
Detailed Data Setting
8.1
Current Output Setting
8-1
This section describes setting of the analog output range.
8.1.1
Oxygen Analyzer - Setting Minimum Current (4 mA)
and Maximum Current (20 mA)
To set the minimum and maximum currents, proceed as follows:
(1) Select [Commissioning] in the Execution/Setup display.
(2) Select the “mA-output setup” in the Commissioning display.
(3) Select “mA-output1” in the mA-outputs display.
(4) Select “Min. oxygen conc” in the mA-output1 range display and press the [Enter] key. The
numeric-data entry display then appears. Enter the oxygen concentration for the minimum
current (4 mA); for example, enter “010” for 10 vol%O2.
(5) Select “Max. oxygen conc” in the mA-output1 range display and enter the oxygen
concentration for the maximum current (20 mA) in the same manner as in step 4 above.
(6) Set mA-output2 in the same way as the setting procedure for mA-output1 given above.
8.1.2
Humidity Analyzer - Setting Minimum Current (4 mA)
and Maximum Current (20 mA)
To set the minimum humidity to 50% H2O and the maximum humidity to 100% H2O,
follow these steps:
(1) Select Setup in the Execution/Setup display.
(2) Select the mA-output setup in the “Commissioning” (Setup) display.
(3) Select mA-output1 in the mA-outputs display.
(4) Select the Max. humidity conc. in the mA-output1 range display and press the [Enter] key.
The numeric-data entry display then appears. Enter the humidity value for the maximum
current (20 mA); for example, enter “100” for 100% H2O.
NOTE
For the humidity measurement, 0% H2O is a default setting for the minimum humidity and 25%
H2O is the default for the maximum humidity. If you first attempt to set 50% H2O for the minimum
humidity, you cannot set it because that value is outside the set range. In such a case, set the
maximum humidity first.
(5) Select the Min. humidity conc. and enter a minimum humidity value, e.g., 050 for 50% H2O.
(6) Follow the above steps to set the mA-output2, if necessary.
8.1.3
Input Ranges
The range low and high values are restricted as follows:
l Oxygen Concentration setting range
The range min. O2 concentration value (corresponding to 4 mA output) can be set to either 0
vol%O2 or in the range of 6 to 76 vol%O2.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
8-2
<8. Detailed Data Setting>
The range max. O2 concentration value (corresponding to 20 mA output) can be set to any value
in the range of 5 to 100 vol%O2, however the range max. setting must be at least 1.3 times the
range min. setting.
If you do not observe this restriction, the measurement will be invalid, and any previous valid
value will be used. The gray area in figure represents the valid setting range.
Setting example 1
If the range minimum (corresponding to 4 mA output) is set to 10 vol%O2 then range maximum
(corresponding to 20 mA output) must be at least 13 vol%O2.
Maximum oxygen concentration, vol%O2 (for a maximum current of 20 mA)
Setting example 2
If the range minimum (corresponding to 4 mA output) is set to 75 vol%O2 then range maximum
(corresponding to 20 mA output) must be at least 75x1.3=98 vol%O2 (rounding decimal part up).
95
85
75
Ranges over which oxygen
concentrations can be set
65
55
45
Outside ranges
35
25
15
5
5
15
25
35
45
55
Minimum oxygen concentration, vol%O2 (for a minimum current of 4 mA)
Minimum-Maximum setting range of oxygen concentration
65
75
F8.0E.ai
Figure A
l Humidity (amount-of-moisture-content) setting range
The minimum humidity is set to 0% H2O or ranges from 26 to 100% H2O. The maximum humidity
ranges from 25% to 100% H2O, and must be greater than 0.8 times plus 23 the humidity set for
the minimum.
Setting example 1
If the setting (for a 4 mA current) is 0% H2O, you must set the maximum (20 mA) point at more
than 25% H2O.
Setting example 2
If the setting (for a 4 mA current) is 26% H2O, you must set the maximum (20 mA) point at more
than 44% H2O, (263 0.8 + 23% H2O). (Numbers after the decimal point are rounded up.)
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
8-3
<8. Detailed Data Setting>
Maximum humidity (for a 20-mA current), % H2O
100
90
Ranges over which oxygen
concentrations can be set
80
70
60
Outside ranges
50
40
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
Minimum humidity (for a 4-mA current), % H2O
F8-2E.ai
Figure B
Max. and Min. Humidity Set Ranges
l “Mixing ratio” setting range
The minimum mixing ratio is set to 0 kg/kg or ranges from 0.201 to 0.625 kg/kg. The maximum
“mixing ratio” setting ranges from 0.2 to 1.0 kg/kg, and must be greater than 1.3 times plus 0.187
the mixing ratio set for the minimum.
Setting example 1
If the setting (for a 4 mA current) is 0 kg/kg, you must set the maximum (20 mA) point at more
than 0.2 kg/kg.
Setting example 2
If the setting (for a 4 mA current) is 0.201 kg/kg, you must set the maximum (20 mA) point at more
than 0.449 kg/kg, (0.201 3 1.3 + 0.187 kg/kg). (Numbers after the decimal point are rounded up.)
1
Ranges over which oxygen
concentrations can be set
0.95
Maximum mixing ratio, kg/kg at 20 mA
0.9
0.85
0.8
0.75
0.7
0.65
0.6
0.55
Outside ranges
0.5
0.45
0.4
0.201
0.25
0.3
0.35
0.4
0.45
0.5
0.55
0.65
F8-3E.ai
Minimum mixing ratio, kg/kg at 4 mA
Figure C
0.6
Max. and Min. Mixing Ratio Set Ranges
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
8-4
<8. Detailed Data Setting>
8.1.4
Setting Output Smoothing Coefficient
When the oxygen concentration in the sample gas fluctuates rapidly, if the measured value is
used for control this can lead to problems with undesirable frequent ON/OFF switching. You can
set a smoothing time constant of between 0 and 255 seconds to reduce the effect. Select the
appropriate output damping constant from the numeric-data entry display. To set 30 seconds,
enter 030.
8.1.5
Selection of Output Mode
You can select whether the relationship between the sample oxygen concentration and the
analog output signal be linear or logarithmic. Press the [Enter] key in the output mode display. A
linear/ logarithmic selection display then appears. Select the desired mode.
NOTE
If you select an output mode of “logarithmic” then regardless of range setting the minimum output
value becomes 0.1 vol%O2 (humidity; 0.1% H2O, mixing ratio; 0.01 kg/kg) fixed.
8.1.6
Default Values
When the analyzer is delivered or reset to defaults, the output current default settings by as
shown in Table 8.1.
Table 8.1
Output Current Default Values
Item
8.2
Default setting
Min. oxygen concentration
0 vol%O2
Max. oxygen concentration
25 vol%O2
Minimum humidity conc.
0 % H2O
Maximum humidity conc.
25 % H2O
Minimum mixing ratio
0.000 kg/kg
Maximum mixing ratio
0.200 kg/kg
Output damping constant
0 (seconds)
Output mode
Linear
Output Hold Setting
The “output hold” functions hold an analog output signal at a preset value during the equipment’s
warm-up time or calibration or if an error arises. Outputs 1 and 2 can not be set individually. Table
8.2 shows the analog outputs that can be retained and the individual states.
Table 8.2
Equipment status
Output
During warm-up
hold values available
4 mA
20 mA
Without hold feature
Retains output from just before
occurrence
Set value (2.4 to 21.6 mA)
○: The output hold functions are available.
○
○
○
Under
maintenance
Under calibration
During blow back
On Error
occurrence(*1)
○
○
○
○
○
○
○
○
○
*1: The output hold functions on error occurrence are unavailable when option code “/C2” or “/C3”(NAMER NE 43 compliant) is
specified.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
8.2.1
<8. Detailed Data Setting>
8-5
Definition of Equipment Status
(1) During warm-up
“During warm-up” is the time required after applying power until sensor temperature stabilizes at
750ºC and the instrument is in the measurement mode.
(2) Under Maintenance
“Under maintenance” is the time that starts when you move to the Execution/Setup display by
touching the [Setup] key on the Basic panel display and ends when you return to the Basic panel
display. It includes when you operates keys on lower level menu displays of the Execution/Setup
display.
(3) Under Calibration (see Chapter 9, Calibration, later in this manual)
For manual Calibration, “Under calibration” is the time that starts when you move to the Manual
calibration display (Figure 7.30) from the Calibration display, lasts while you are operating keys
for performing calibration manually, and ends when you press the End key and after a preset hold
time has elapsed.
For semi-automatic calibration, “Under calibration” is the time required from entering calibration
instructions to perform a, either by using the touchpanel or by a contact input, calibration until the
hold (output stabilization) time elapses.
For automatic calibration, “Under calibration” is the time required, after performing an appropriate
calibration until the hold (output stabilization) time elapses.
(4) During Blow back (see Section 10.4, Blow Back, later in this manual)
During semi-automatic blow back:
“During semi-automatic blow back” is the time required after pressing the [Blow back start] key,
by using the touchpanel or entering a blow back start instruction by using a contact input, until the
blow back time and hold (output stabilization) time elapse.
During automatic blow back:
“During automatic blow back” is the time required after reaching the blow back start time until the
blow back time and hold (output stabilization) time elapse.
(5) On Error occurrence
This is the time at which any of Errors 1 to 4 occurs.
8.2.2
Preference Order of Output Hold Value
The output hold value takes the following preference order:
On error occurrence
Preference order (high)
Under calibration or during blow back
Under maintenance
During warm-up
8.2.2.siki
For example, if the output current is set to 4 mA during maintenance, and no output-hold output
for during calibration is preset, the output is held at 4 mA during the maintenance display.
However, the output hold is released at the time of starting the calibration, and the output will be
again held at 4 mA after completing the calibration and when the hold (output stabilization) time
elapses.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
8.2.3
8-6
<8. Detailed Data Setting>
Output Hold Setting
To set the output hold, follow these steps:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
Then select “Commissioning” in the Execution/Setup display. Next, select the “mA-output
setup” and then the “mA-outputs presets” display as shown in Figure 8.2.
mA-outputs presets
Warm up:
4mA
Preset value: 4 . 0 mA
Maintenance: Hold
Preset value: 4 . 0 mA
Cal.blow back: Hold
Preset value: 4 . 0 mA
Error:
Preset
Preset value: 3 . 4 mA
mA-outputs presets
Enter
F8.1E.ai
Figure 8.1
mA-outputs presets Display
Warm up:
4mA
Preset value: 4 . 0 mA
Maintenance: Hold
Non-Hold
Preset value: Hold
4.0mA
Cal.blow back: Preset
Hold
Preset value: 4 . 0 mA
Error:
Preset
Preset value: 3 . 4 mA
Enter
F8.2E.aiS
Figure 8.2
Selecting Maintenance
(2) From this display (Fig. 8.2), select the desired display. Figure 8.3 shows an example
of selecting “Maintenance”. Select the desired output status.
(3) If a preset value is selected, set the corresponding output current. If you select a preset
value just below Maintenance on the screen, the numeric-data entry display will appear.
Enter the current value you want. To set 10 mA, type in 010 and press the [Enter] key to
complete the setting. The setting range is from 2.4 to 21.6 mA.
NOTE
“Error” of mA-outputs presets is not displayed when option code “/C2” or “/C3” (NAMUR NE 43
compliant) is specified.
8.2.4
Default Values
When the analyzer is delivered, or if data are initialized, output hold is the default as shown in
Table 8.3.
Table 8.3
Output Hold Default Values
Status
8.3
Output hold setting
Preset value
During warm-up
4 mA
4 mA
Under maintenance
Holds output at value just before maintenance started.
4 mA
Under calibration or blow back
Holds output at value just before starting calibration or
blow back.
4 mA
On Error occurrence
Holds output at a preset value.
3.4 mA
Alarms Setting
The analyzer enables the setting of four alarms — high-high, high, low, and low-low alarms
— depending upon the measurement conditions. The following section sets out the alarm
operations and setting procedures.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
8-7
<8. Detailed Data Setting>
8.3.1
Setting the Alarm Values
(1) High-high and high alarm values
If high-high and high alarm values are set to ON, then alarms occur if measured valves exceed
the alarm set values.
(2) Low and low-low alarm values
If low-low and low alarm values are set, then alarms occur if measured values fall below the
alarm set values.
8.3.2
Alarm Output Actions
If the measured values of the oxygen concentration fluctuate between normal (steady state)
values and the alarm setting, alarm outputs may be frequently issued and canceled.
To avoid this, set the alarm delay and hysteresis for alarm canceling under the alarm output
conditions, as Figure 8.3 shows.
When a delay time is set, an alarm will not be issued so quickly even if the measured value differs
from the steady state and enters the alarm setpoint range.
If the measured value remains within the alarm setpoint range for a certain period of time (for the
preset delay time), an alarm will be issued. On the other hand, there will be a similar delay each
time the measured value returns to the steady state from the alarm setpoint range (canceling the
alarm status).
If hysteresis is set, alarms will be canceled when the measured value is less than or greater than
the preset hysteresis values.
If both the delay time and hysteresis are set, an alarm will be issued if the measured value is in
the alarm setpoint range and the delay time has elapsed.
For the alarm to be reset (canceled), the measured value must be beyond the preset hysteresis
value and the preset delay time must have elapsed.
Refer to Figure 8.3 for any further alarm output actions. The delayed time and hysteresis settings
are common to all alarm points.
Alarm range
A
B
C
D
7.5%
High limit alarm setpoint
Hysteresis
2.0%
5.5%
Oxygen concentration
Delay time:
Delay time:
5 seconds
5 seconds
Delay time:
5 seconds
Alarm output ON
OFF
Figure 8.3
F8.3E.ai
Alarm Output Action
In the example in Figure 8.3, the high limit alarm point is set to 7.5 vol%O2, the delay time is set to
five seconds, and hysteresis is set to 2 vol%O2.
Alarm output actions in each section in this figure are as follows:
A. Although the oxygen concentration value exceeds the high limit alarm setpoint, it falls below
the high limit alarm setpoint before the preset delay time of five seconds elapses. So, no
alarm is issued.
B. The oxygen concentration value exceeds the high limit alarm setpoint and the delay time
elapses during that measurement. So, an alarm is issued.
C. Although the oxygen concentration value falls below the hysteresis set value, the value rises
again and exceeds the hysteresis set value before the preset delay time elapses. So, the
alarm is not canceled.
D. The oxygen concentration value falls below the hysteresis set value and the preset delay
time elapses, so the alarm is canceled.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
8.3.3
8-8
<8. Detailed Data Setting>
Alarm Setting Procedure
To set the alarm setpoints, follow these steps:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
(2) Select “Commissioning” in the Execution/Setup display. The Commissioning (Setup) display
then appears.
(3) Select the “Alarms setup” in the Commissioning (Setup) display. The Alarms setup display
shown in Figure 8.4 then appears.
• To set the hysteresis, proceed to the following steps.
(4) Select “Hysteresis” in the Alarms setup display. The numeric-data entry display then
appears. Enter the desired hysteresis value as a percent of oxygen concentration. To set 2.5
vol%O2, enter “0025.” The hysteresis setting can be in the range of 0 to 9.9 vol%O2.
• To set the delay time, proceed as per the following steps.
(5) Select the “Contact delay” in the Alarms setup display. The numeric-data entry display then
appears. Enter the desired delay time, in seconds. To set three seconds, enter “003.” The
delay time setting can be in the range of 0 to 255 seconds.
• To set the alarm point, proceed to the following steps.
(6) Select the “Setpoints” in the Alarms setup display. The Oxygen alarms display then appears,
as shown in Figure 8.5.
(7) When you select “High alarm” in the Oxygen alarms display, the “OFF” or “ON” selection
display then appears. If you select “ON,” the High alarm will then be enabled (enable/
disable).
(8) To set the High alarm values select “Set value” just below the High alarm. The numeric-data
entry display then appears. Enter the alarm set value (percent of oxygen concentration). If
you want to set the alarm value to 10 vol%O2, enter “010.”
(9) Set the other alarm settings in the same manner as in the steps above.
NOTE
No alarm is issued when alarm is set to “OFF” (disabled). To use the alarm functions, be sure to
set the alarms “ON”.
Oxygen alarms
Alarms setup
Parameter:
Hysteresis:
Oxygen
0.1%O2
Contact delay:
3s
Setpoints
Enter
F8.4E.ai
Figure 8.4
Alarms Setup Display
High High alarm: OFF
Set value: 1 0 0 . 0 % O2
High alarm:
ON
Set value: 1 0 0 . 0 % O2
Low alarm:
OFF
Set value:
0 . 0 % O2
Low Low alarm: OFF
Set value:
0 . 0 % O2
Enter
F8.5E.ai
Figure 8.5
Oxygen Alarms Display
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
8.3.4
8-9
<8. Detailed Data Setting>
Default Values
When the analyzer is delivered, or if data are initialized, the default alarm set values are as
shown in Table 8.4.
Table 8.4
Alarm Setting Default Values
Set item
8.4
Oxygen concentration
Humidity (amount of
moisture content)
Mixing ratio
Setting
range
Default
setting
Setting
range
Default
setting
Setting
range
Default
setting
Hysteresis
0 to 9.9
vol%O2
0.1 vol%O2
0 to 9.9
%H2O
0.1 %H2O
0 to 0.1 kg/
kg
0.001 kg/kg
Delay time
0 to 255
seconds
3 seconds
0 to 255
seconds
3 seconds
0 to 255
seconds
3 seconds
High-high limit
alarm
–
OFF
–
OFF
–
OFF
High-high limit
alarm setpoint
0 to 100
vol%O2
100 vol%O2
0 to 100
%H2O
100.0
%H2O
0 to 1 kg/kg
1 kg/kg
High limit alarm
–
OFF
–
OFF
–
OFF
High limit alarm
setpoints
0 to 100
vol%O2
100 vol%O2
0 to 100
%H2O
100.0
%H2O
0 to 1 kg/kg
1 kg/kg
Low limit alarm
–
OFF
–
OFF
–
OFF
Low limit alarm
setpoint
0 to 100
vol%O2
0 vol%O2
0 to 100
%H2O
0.0 %H2O
0 to 1 kg/kg
0 kg/kg
Low-low limit alarm
–
OFF
–
OFF
–
OFF
Low-low limit alarm
setpoint
0 to 100
vol%O2
0 vol%O2
0 to 100
%H2O
0.0 %H2O
0 to 1 kg/kg
0 kg/kg
Output Contact Setup
8.4.1
Output Contact
Mechanical relays provide contact outputs. Be sure to observe relay contact ratings. (For details,
see Section 2.1, General Specifications) The operation modes of each contact output are as
follows. For output contacts 1 to 3 you can select open or closed contact when the contact is
“operated”. Default is closed. For output contact 4, contact is closed.
When power fails, contact outputs 1 to 3 are open, and 4 is closed.
Table 8.5
State when contact “ operated ”
When no power is applied
to this equipment
Output contact 1
Open (deenergized) or closed (energized) selectable.
Open
Output contact 2
Open (deenergized) or closed (energized) selectable.
Open
Output contact 3
Open (deenergized) or closed (energized) selectable.
Open
Output contact 4
Closed (deenergized) only
Closed
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
8-10
<8. Detailed Data Setting>
8.4.2
Setting Procedure
To set the output contact, follow these steps.
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
(2) Select “Commissioning” in the Execution/Setup display. The Commissioning (Setup) display
then appears.
(3) Select the “Contact setup” in the Commissioning (Setup) display. The Contact setup display
shown in Figure 8.6 then appears.
(4) Select the desired output contact. This section shows an example where “Output contact1”
is selected (see Figure 8.7).
(5) Each set item and the selected items are briefly described in Table 8.6. The following
describes an example of setting where Output contact 1 is closed during calibration.
(6) Select “Others” in the Output contact 1 display. The “Contact1 Others” display shown in
Figure 8.8 then appears. Select “Calibration” in the Contact1 Others display.
(7) The ON or OFF selection display then appears. Select “ON” herein.
(8) Press the [Reject] key to go back to the previous display.
(9) Move the pointer to “During power-off the contact is open and in condition it is Open” and
press the [Enter] key. The “OFF” or “ON” selection display then appears. If you select “OFF,”
this means “Open” in normal conditions and “Closed” when the contact output is on.
CAUTION
The contact output 4 is fixed as “ close in power ON”, which cannot be changed by setting.
Output contact 1
Contact setup
Output contact 1
Output contact 2
Alarms
Others
During power-off the
contact is open and
in condition it is Open
Output contact 3
lnput contacts
Enter
Enter
F8.7E.ai
Figure 8.6 Contact Setup Display
Contact1
F8.8E.ai
Figure 8.7 Output Contact 1 Display
Others
Warm up:
Range change:
Calibration:
Maintenance:
Blow back:
Temp.input high:
Cal.gas press.low:
Process up set:
ON
OFF
OFF
ON
OFF
OFF
OFF
OFF
Enter
F8.9E.ai
Figure 8.8
“Contact1 Others” Display
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<8. Detailed Data Setting>
Table 8.6
Output Contact Settings
Item to be selected
Alarm
and
Error
settings
Other
settings
8-11
Brief description
High-high limit alarm
If “High-High alarm ON” is selected, contact output occurs when the
high-high limit is issued. To do this, it is required, in Alarms setup, that
the high-high alarm be set on beforehand (see Section 8.3).
High limit alarm
If “High alarm ON” is selected, contact output occurs when the high
limit alarm is provided. To do this, it is required, in Alarms setup, that
the high limit alarm be set on beforehand (see Section 8.3).
Low limit alarm
If “Low alarm ON” is selected, contact output occurs when the low
limit alarm is provided. To do this, it is required, in Alarms setup, that
the low limit alarm be set on beforehand (see Section 8.3).
Low-low limit alarm
If “Low-Low alarm ON” is selected, contact output occurs when the
low-low limit alarm is issued. To do this, it is required, in Alarms setup,
that the low-low alarm be set on beforehand (see Section 8.3).
Calibration coefficient alarm
If Calibration coefficient alarm is ON (enabled), then when a zero
calibration coefficient alarm (Alarm 6) or span calibration coefficient
alarm (Alarm 7) occurs then calibration coefficient alarm contact
output occurs (see Sec. 12.2.1)
Startup power stabilization
timeout alarm
If set ON then contact output occurs when startup power stabilization
timeout alarm (Alarm 8) occurs (see Sec.12.2.1)
Error
If “Error ON” is selected, contact output occurs when an error is
issued. (See Chapter 12, Troubleshooting).
Warm-up
If “Warm-up ON” is selected, contact output occurs during warm-up.
For the definition of Warm-up (see Section 8.2.1).
Output range change
If “Range change ON” is selected, contact output occurs (“answerback signal to a range change signal”) while a range change signal is
applied to a contact input. To do this, it is required, in Input contacts
setup, that the range change be selected beforehand. For more on
this (see Section 8.5).
Calibration
If “Calibration ON” is selected, contact output occurs during
calibration. For the definition of Under calibration (see Section 8.2.1).
Maintenance
If “Maintenance ON” is selected, contact output occurs during
maintenance. For the definition of Under maintenance (see Section
8.2.1).
Blow back
If “Blow back ON” is selected, contact output occurs during blow
back. For the definition of During blow back (see Section 8.2.1).
High limit temperature alarm Not supported by the oxygen analyzer.
Calibration gas press. low
If “Cal. gas press. low ON” is selected, contact output occurs
(“answer-back signal to a calibration gas low pressure signal”)when
a calibration gas low pressure signal is applied to the contact input.
To do this, it is required, in Input contacts setup, that “Cal. gas press.
low” be selected beforehand. For more on this (see Section 8.5).
Process upset
If “Process upset” is selected, contact output occurs (“answer-back
signal to a process upset signal) when the process upset signal is
applied to the contact input. To do this, it is required, in Input contacts
setup, that “Process upset” be selected beforehand (see Section 8.5).
Note: To provide an alarm with an output contact, be sure to make an alarm setting.
When using contact output as an answer-back signal for an input contact, be sure to make an input contact setting.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
8.4.3
8-12
<8. Detailed Data Setting>
Default Values
When the analyzer is delivered, or if data are initialized, alarm and other setting defaults are as
shown in Table 8.7.
Table 8.7
Output Contact Default Settings
Item to be selected
Output
contact 1
Output
contact 2
Output
contact 3
Output
contact 4
High-high limit alarm
Alarm
settings
�
High limit alarm
ON
�
Low limit alarm
ON
�
Low-low limit alarm
�
Calibration coefficient alarm
�
Startup power stabilization timeout alarm
�
Error
ON
Warm-up
ON
�
Output range change
�
Calibration
Other
settings
ON
Maintenance
�
ON
�
Blow back
�
High limit temperature alarm
Calibration gas press. low
�
(default)
(default)
(default)
�
Process upset
�
Operating contact status
Open
Closed
Closed
(fixed)
Closed
Note: Blank boxes in the above table indicate that the default is “disabled“.
8.5
Input Contact Settings
8.5.1
Input Contact Functions
The converter input contacts execute set functions by accepting a remote dry-contact (“voltagefree contact”) signal. Table 8.8 shows the functions executed by a remote contact signal.
Table 8.8
Input Contact Functions
tem
Function
Calibration gas pressure low
Contact input disables Semi-automatic or Automatic Calibration.
Measuring range change
While contact input is On, range of Analog Output 1 is switched as follows:
When analog output 1 range is set to “Oxygen”, then range is switched to 0 to
25% O2. When analog output 1 range is set to “Humidity”, then output range is
switched to 0 to 100% H2O. When analog output 1 range is set to “Mixing ratio”,
then output range is switched to 0 to 1 kg/kg.
While range is switched by the contact input, [Range] is displayed on the
screen. See Figure 8.9.
Calibration start
Contact input starts Semi-Automatic Calibration. Calibration Mode setting must
be [Semi_Auto] or [Auto]. Contact signal must be applied for at least 1 sec. Even
if input signal continues to be applied, calibration is not repeated unless contact
input is released then reapplied.
Process upset
(Combustible gas detection)
If the Contact signal is ON, heater power will be switched off. (A one- to
11-second time interval single-output signal is available as a contact signal.) If
this operation starts, the sensor temperature decreases and an error occurs. To
restore it to normal, turn the power off and then back on, or reset the analyzer.
Blow back start
Contact input starts Blow back. Contact signal must be applied for at least 1 sec.
Even if input signal continues to be applied, Blow back is not repeated unless
contact input is released then reapplied. (Refer to Section 10.4, Blow Back.)
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
8-13
<8. Detailed Data Setting>
Tag:
21.0
%O2
7.35mA -Output1
7.35mA -Output2
Range
F8.9E.ai
Figure 8.9
Changing Measuring Range with Input Contact
CAUTION
• Measurement range switching function by an external contact input is available for analog
output1 only.
• When making a semi-automatic calibration, be sure to set the semi-automatic or automatic
mode using the Calibration setup display.
• When carrying out “Blow back,” be sure to set “Blow back” in the output contact setup.
• When the combustible gas detection signal is sent to the contact input, the converter will cut
the power supply to the heater of the detector. As a result, the heater temperature becomes
low and Error 1 or Error 2 happens.
8.5.2
Setting Procedure
This setting example shows how to set “When contact input 1 Opens, Start Semi_Auto
Calibration”.
(1) From the Basic panel display touch [Setup] key, and the [Execution/Setup] display appears.
(2) Select “Commissioning” and the Commissioning display appears.
(3) Select “Contact Setup” then “Input contacts”.
(4) Select “Input 1” . Function Selection window (Fig 8.11) appears.
(5) Select “Calibration start”.
(6) For Input contact 1 select (operate if) “Closed”. (You can select “Open” or “Closed”).
(7) Change Closed to Open.
Input contacts
Input contacts
Input1: No function
Input1: No
No function
function
Cal.
gas
Input1 contact
is press.
active low
at:
Range change
Closed start
Calibration
up set
Input2: Process
No function
Blow back start
Input2 contact is active at:
Closed
Input1 contact is active at:
Closed
Input2: No function
Input2 contact is active at:
Closed
Enter
F8.10E.ai
Figure 8.10
Input Contacts Display
Enter
F8.11.ai
Figure 8.11
Input Contacts Display
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
8-14
<8. Detailed Data Setting>
8.5.3
Default Values
All contact inputs are set to “No function”(disabled) prior factory shipment or after data
initialization..
8.6
Other Settings
8.6.1
Setting the Date-and-Time
The following describe how to set the date-and-time. Automatic calibration or blow back works
following this setting.
Proceed as follows:
(1) From the Basic panel display if you touch the [Setup] key, the Execution/Setup display
appears.
(2) Select “Commissioning” and the Commissioning display appears.
(3) Select “Others” and the display of Fig 8.12 appears.
(4) Select “Clock” and the display of Fig 8.13 appears.
(5) Select “Set date” and the numerical entry display appears. To enter June 21. 2004 , enter
210604. Touch the [Enter] key and you revert to the display of Fig 8.13.
(6) Select “Set time”. Enter the time in 24-hour format. To set 2:30 pm, enter 1430 on the
numerical display and touch enter. Touch the [Enter] key. The clock starts from 00 seconds..
Others
Clock
Averaging
Fuel setup
Purging
Unit setup
Passwords
Defaults
Clock
Set date: 1 3 / 0 7 / 0 0
Set time: 1 4 : 3 9
1 3 J ul. 2 0 0 0
14:39:55
Enter
F8.12E.ai
Figure 8.12
8.6.2
Other Settings
Enter
F8.13E.ai
Figure 8.13
Clock Display
Setting Periods over which Average Values Are
Calculated and Periods over which Maximum and
Minimum Values Are Monitored
The equipment enables the display of oxygen concentration average values and maximum and
minimum values under measurement (see Section 10.1, later in this manual).
The following section describes how to set the periods over which oxygen concentration average
values are calculated and maximum and minimum values are monitored.
Procedure
(1) From the Basic panel display touch [Setup] key, and the Execution/Setup display appears.
(2) Select “Commissioning” and the Commissioning display appears.
(3) Select “Others” then Averaging. The display of Fig. 8.14 appears.
(4) Select “Set period over which average is calculated” and numerical entry display appears.
To set 3 hours, enter 003. Entry range is 1 to 255 hours.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<8. Detailed Data Setting>
8-15
(5) Select “Set period over which maximum and minimum is stored” and numerical entry display
appears. To set 48 hours, enter 048. Entry range is 1 to 255 hours.
Averaging
Set period over which
average is calculated:
1h
Set period over which
maximum and minimum is
stored:
24h
Enter
F8.14E.ai
Figure 8.14 Setting Average-Value Calculation Periods and
Maximum- and Minimum-Value Monitoring Periods
Default Values
“Set period over which average is calculated” is set to 1hr, and “Set period over which maximum
and minimum is stored” is set to 24 hrs prior factory shipment or after data initialization.. .
8.6.3
Setting Fuels
Input Parameters
The analyzer calculates the moisture content contained in exhaust gases. The following sets
forth the fuel parameters necessary for calculation and their entries. The moisture quantity may
be mathematically expressed by:
(water vapor caused by combustion and water vapor contained in the exhaust gas)
Moisture quantity =
+ (water vapor contained in air for combustion)
actual exhaust gas (including water vapor) per fuel
x 100 ..... Equation 1
= (Gw + Gw1)/G x 100
=
Gw + (1.61 x Z x m x Ao)
Go + Gw + [(m - 1) Ao + (1.61 x Z x m x Ao)]
Gw + (1.61 x Z x m x Ao )
X + Ao x m
x 100
............ Equation 2
x 100
where,
Ao : Theoretical amount of air per unit quantity of fuel, m3/kg (or m3/m3) ... 2 in Table 8.8
G: Actual amount of exhaust gas (including water vapor) per unit quantity of fuel, m3/kg (or m3/m3)
Gw : Water vapor contained in exhaust gas per unit quantity of fuel (by hydrogen
............ 1 in Table 8.8
and moisture content in fuel), m3 /kg (or m3/m3)
Gw1: Water vapor contained in exhaust gas per unit quantity of fuel (moisture content
in air), m3 /kg (or m3/m3)
Go: Theoretical amount of dry exhaust gas per unit quantity of fuel, m3/kg (or m3/m3)
m: Air ratio
X : Fuel coefficient determined depending on low calorific power of fuel, m3/kg (or m3/m3) ....... 3 in Table 8.8
Z : Absolute humidity of the atmosphere, kg/kg ....... Figure 8.16
Fill in the boxes with fuel parameters in Equation 2 above to calculate the moisture content.
Use Ao, Gw and X shown in Table 8.9. If there are no appropriate fuel data in Table 8.9, use the
following equations for calculation. Find the value of “Z” in Equations 1 and 2 using Japanese
Standard JIS B 8222. If a precise measurement is not required, obtain the value of “Z” using a
graph for the absolute humidity indicated by a dry and wet bulb hygrometer.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<8. Detailed Data Setting>
8-16
For liquid fuel
Amount of water vapor in exhaust gas (Gw) = (1/100) {1.24 (9h + w)} (m3 /kg)
(m3 /kg)
Theoretical amount of air (Ao) = 12.38 x (Hl/10000) – 1.36
Low calorific power = Hl
X value = (3.37 / 10000) x Hx – 2.55 (m 3 /kg)
where,
Hl: low calorific power of fuel
h: Hydrogen in fuel (weight percentage)
w: Moisture content in fuel (weight percentage)
Hx: Same as numeric value of Hl
For gas fuel
Amount of water vapor in exhaust gas (Gw) = (1/100) {(h2) + 1/2 ∑y (Cx Hy) + wv} (m3 /m 3 )
Theoretical amount of air (Ao) = 11.2 x (Hl/10000)
(m3 /m3 )
Low calorific power = Hl
X value = (1.05 / 10000) x Hx
where,
3
3
(m /m )
Hl: low calorific power of fuel
CxHy: Each hydrocarbon in fuel (weight percentage)
h2: Hydrogen in fuel (weight percentage)
wv: Moisture content in fuel (weight percentage)
Hx: Same as numeric value of Hl
For solid fuel
Amount of water vapor in exhaust gas (Gw) = (1/100) {1.24 (9h + w)} (m3 /kg)
Theoretical amount of air (Ao) = 1.01 x (Hl / 1000) + 0.56
3
(m /kg)
Low calorific power = Hl = Hh – 25 (9h + w) (kJ/kg)
X value = 1.11 - (0.106 / 1000 ) x Hx
where,
3
3
(m /m )
w: Total moisture content in use (weight percentage)
h: Hydrogen content (weight percentage)
The average hydrogen content of coal mined in Japan, which is a dry ash-free type, is
5.7 percent. Accordingly, "h" may be expressed mathematically by:
h = 5.7 [{100 – (w + a)} / 100] x (100 – w) / (100 – w1)
where,
a: Ash content (%)
w1: Moisture content (%), analyzed on a constant humidity basis
Hh: Higher calorific power of fuel (kJ/kg)
Hl: Low calorific power of fuel (kJ/kg)
Hx: Same numeric value of Hl
Figure 8.15
Calculation Formula
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
8-17
<8. Detailed Data Setting>
40
39
0.046
38
0.044
0.042
37
0.040
36
0.038
35
0.036
34
0.034
33
0.032
32
0.030
31
0.028
30
Wet-bulb
temperature, °C
29
0.026 Absolute
28
0.024
27
26
0.022
25
0.020
24
0.018
22
20
0.016
18
0.014
16
0.012
14
12
4
2
0
8
6
humidity, kg/kg
0.010
10
0.008
0.006
0.004
-2
0.002
0
2
4
6
8
10
12 14 16
18 20
22 24 26 28 30 32 34 36
Dry-bulb temperature, °C
Figure 8.16
38 40
0.000
F8.16E.ai
Absolute Humidity of Air
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
8-18
<8. Detailed Data Setting>
Table 8.9
Fuel Data
• For liquid fuel
Fuel
properties Specific
weight
kg/l
Type
Chemical component
(weight percentage)
C
H
O
N
S
w
Calorific power
kJ/kg
Ash Higher Lower
content order order
Theoretical
amount of
air for
combustion
Nm3/kg
Amount of combustion
gas Nm3/kg
X
value
N2 Total
CO2 H2O SO2
Kerosene
0.78~
85.7 14.0
0.83
0.5
0.0
0.0
46465 43535
11.4
1.59 1.56 0.00 9.02 12.17 0.96
Light oil
0.81~
85.6 13.2
0.84
1.2
0.0
0.0
45879 43032
11.2
1.59 1.47 0.00 8.87 11.93 0.91
0.5
0.5
0.3 0.05 45544 42739
10.9
1.60 1.34 0.00 8.61 11.55 0.89
0.5
2.0
0.4 0.05 45125 42320
10.8
1.58 1.32 0.01 8.53 11.44 0.86
0.4
3.0
0.5 0.05 43827 41274
10.7
1.58 1.27 0.02 8.44 11.31 0.77
0.4
1.5
0.5
0.1
43952 41441
10.7
1.61 1.22 0.01 8.43 11.27 0.79
0.5
0.4
3.5
0.5
0.1
43116 40646
10.5
1.58 1.20 0.02 8.32 11.12 0.72
0.5
0.4
1.5
0.6
0.1 43660 41190
10.7
1.61 1.22 0.01 8.43 11.27 0.77
A
0.85~
85.9 12.0 0.7
Heavy No.1
0.88
oil
class
No.2 0.83~
84.6 11.8 0.7
1
0.89
B
Heavy oil
class 2
0.90~
84.5 11.3 0.4
0.93
No.1 0.93~
0.95
86.1 10.9 0.5
Heavy
oil
0.94~
84.4 10.7
C class No.2
0.96
3
No.3 0.92~
86.1 10.9
1.00
• For gas fuel
3
Theoretical amount of air
Fuel
properties Specific
Type
1
2
weight
kg/Nm3
Chemical component
(weight percentage)
CO
H2 CO2 CH4 CmHn O2
Coke oven
gas
0.544
9.0 50.5 2.6 25.9 3.9
Blast furnace
gas
1.369
25.0 2.0 20.0
2.0 88.4 3.2
0.1
N2
8.0
53.0
1.6
Theoretical
amount of
air for
combustion
Lower Nm3/m3
Combustion
product,
Nm3/ m3
Calorific power
kJ/Nm3
4.2
Higher
order
order
20428 18209
CO2 H2O N2
X
value
Total
4.455
0.45 1.10 3.60 5.15
0.46
3349
0.603
0.45 0.02 1.01 1.48
0.08
37883 34074
9.015
0.98 1.88 7.17 10.03 0.86
3391
Natural gas
0.796
Propane
2.030
C3H8 90%, C4H10 10%
102055 93976
24.63
3.10 4.10 19.5 26.7
2.36
Butane
2.530
C3H8 10%, C4H10 90%
125496 115868
30.37
3.90 4.90 24.0 32.8
2.91
1.0 1.89 2.89
0.27
(Molecular Formula)
(Gases)
Oxygen
1.43
O2
Nitrogen
1.25
N2
Hydrogen
0.09
H2
12767 10758
2.390
Carbon
monoxide
1.25
CO
12642 12642
2.390
1.0
1.89 2.89
0.32
Carbon dioxide 1.96
CO2
Methane
0.72
CH4
39750 35820
9.570
1.0
2.0 7.57 10.6
0.90
Ethane
1.34
C 2H 6
69638 63744
16.74
2.0
3.0 13.2 18.2
1.60
Ethylene
1.25
C 2H 4
62991 59060
14.35
2.0
2.0 11.4
15.4
1.48
Propane
1.97
C 3H 8
99070 91255
23.91
3.0
4.0 18.9 25.9
2.29
2.59
C4H10
128452 118623
31.09
4.0
5.0 24.6 33.6
2.98
Butane
2
1
3
T8.8E.ai
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
8-19
<8. Detailed Data Setting>
Procedure
To make a fuel setting, follow these steps:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
(2) Select “Commissioning” in the Execution/Setup display. The Commissioning (Setup) display
then appears.
(3) Select “Others” in that display and then the Fuel setup shown in Figure 8.17.
(4) Choose the “Theoretical air quantity required” and the “Contents of moisture in exhaust gas”
in turn. The numeric-data entry display then appears. Enter numeric data using the numeric
keys.
(5) Choose “more” in the Fuel setup display. The Fuel setup shown in Figure 8.18 then appears.
(6) Set the numeric data to the “Value of the X coefficient” and then to the “Absolute humidity of
the atmosphere”.
(7) Choose “finished” to return to the display shown in Figure 8.17.
Fuel setup
Fuel setup
Value of the X coefficient:
1.00
Absolute humidity of the
atmosphere:
0. 1 0 0 0 kg/kg
finished.
See Manual for details!
Contents of moisture in
exhaust gas: 1. 0 0 m3/kg
Theoretical air quantity
required:
1. 0 0 m3/kg
more.....
Enter
Enter
F8.17E.ai
Figure 8.17
Fuel Setup
F8.18E.ai
Figure 8.18
Fuel Setup
Default Values
When the analyzer is delivered, or if data are initialized, default, parameter settings are as shown
in Table 8.10.
Table 8.10
Default Settings for Fuel Values
Item
8.6.4
Default setting
Amount of water vapor in exhaust gas
0.00 m3/kg (m3)
Theoretical amount of air
1.00 m3/kg (m3)
X value
1.00
Absolute humidity of the atmosphere
0.0 kg/kg
Setting Measurement Gas Temperature and Pressure
The analyzer calculates the moisture content contained in exhaust gases and saturated water
vapors from the entered gas temperature and pressure to obtain the relative humidity and dew
point. The relative humidity may be obtained using the following theoretical equation (JIS Z
8806).
To obtain the relative humidity:
The relative humidity U that is obtained from JIS Z 8806 is:
U = e/es x 100
where, e = water vapor pressure of moist air
es = Saturated water vapor
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<8. Detailed Data Setting>
8-20
Since the gas-pressure ratio is equal to the volume ratio, the above equation may be expressed
mathematically by:
U = P x H/ es x 100
where, P = Gas pressure
H = moisture content (volume ratio)
The saturated water vapor pressure es is determined by a gas temperature, so the relative
humidity can be obtained by entering the above parameters.
To obtain the dew point:
The dew point is the temperature at which a water vapor pressure in the moist air is equal to the
saturated water vapor pressure.
The water vapor pressure in the moist air can be obtained from the gas pressure ad volume ratio
(= pressure ratio), as given below.
e=PxH
where, e = water vapor pressure in moist air
P = gas pressure
H = Humidity (moisture content) (volume ratio)
Use the above equation to find the water vapor in the moist air, and use the theoretical equation
(JIS Z 8806) to obtain the temperature at which that water vapor is equal to the saturated water
vapor pressure.
l Setting Measurement Gas Temperature
There are two ways of entering measurement gas temperatures: one is to measure actual gas
temperature using a two-wire temperature transmitter and the other is to enter the preset value
manually.
Set the measurement gas temperature as follows:
(1) Press the Setup key in the basic panel display to display the Execution/Setup display.
(2) Select Setup in the Execution/Setup display. The “Commissioning” (Setup) display then
appears.
(3) Select Others in that display and then the Exhaust gas setup shown in Figure 8.19.
(4) Point to the Temperature input mode and press the [Enter] key. Then the display for
selecting either “Preset” or “mA input,” appears. Choose the desired one for your system
requirements.
(5) If you choose the “Preset,” a display for entering numeric data then appears. Enter the
measurement gas temperature (see CAUTION).
(6) If you choose “mA-input,” the Exhaust gas setup display shown in Figure 8.20 appears.
Enter temperatures at 4- and 20-mA outputs of the temperature transmitter from the numeric
data entry display (See CAUTION).
(7) If you use measurement gas temperature alarms, choose the “Alarm value of temperature”
and enter an alarm temperature from the numeric data entry display.
NOTE
• The critical temperature of the saturated water vapor pressure is 3748 C. If a gas
temperature exceeding 3708 C is entered, no correct calculation will be obtained.
• If an invalid value is set, no correct calculation will be obtained. Be sure to check allowable
temperature ranges of the temperature transmitter you use, and then enter the value
correctly.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
8-21
<8. Detailed Data Setting>
Exhaust gas setup
Exhaust gas setup
Temperature input mode:
mA-input
Input temperature value:
3 0 0 ºC
1 0 1. 3 3 kPa
Pressure:
Enter
F8-20E.ai
Figure 8.19
Temperature input mode:
mA-input
Temperature at 4mA:
0 ºC
Temperature at 20mA:
1 0 0 0 ºC
Alarm value of
temperature: 1 0 0 0 ºC
Pressure: 1 0 1 . 3 3 kPa
Enter
F8-21E.ai
Exhaust Gas Setup
Figure 8.20
Display (Preset Entry)
Exhaust Gas Setup Display
(Entering mA-input)
l Setting Measurement Gas Pressure
To set the measurement gas pressure, follow these steps:
(1) Call up the Exhaust gas setup display and choose Pressure.
(2) Enter the measurement gas pressure (absolute pressure) from the numeric data entry display.
l Default Values
When the analyzer is delivered, or if data are initialized, measurement gas temperature and
pressure set ranges and their default settings are as shown in Table 8.11.
Table 8.11
Measurement Gas Temperature and Pressure Set Ranges and Default Settings
Set Item
Set range
Temperature input selection
Default setting
Preset value
Measurement gas temperature
0 to 3000°C
300°C
Temperature at 4 mA
0 to 3000°C
0°C
Temperature at 20 mA
0 to 3000°C
1000°C
Absolute humidity of the atmosphere
0 to 689.47 kPa abs 101.33 kPa abs
8.6.5
Setting Purging
Purging is to remove condensed water in the calibration gas pipe by supplying a span calibration
gas for a given length of time before warm-up of the detector. This prevents cell breakage during
calibration due to condensed water in the pipe.
Open the solenoid valve for the automatic calibration span gas during purging and after the purge
time has elapsed, close the valve to start warm-up.
Purging is enabled when the cell temperature is 100°C or below upon power up and the purge
time is set in the range of 1 to 60 minutes.
Tag:
30
°C
4mA -Output1
4mA -Output2
Purge
Hold
F8.19.ai
Figure 8.21
Display during Purging
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
8-22
<8. Detailed Data Setting>
Procedure
Set the purging time as follows:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
(2) Select “Commissioning” in the Execution/Setup display. The Commissioning (Setup) display
then appears.
(3) Select “Others” in that display and the Others display then appears, as shown in Figure 8.22.
(4) Select “Purging”. The purging time setting display appears, as shown in Figure 8.23.
(5) Point to the “Purging time” and press the [Enter] key. Then the display for selecting purging
time appears.
(6) Enter the desired numeric value from the numeric-data entry display.
The allowable input ranges from 0 to 60 minutes.
Purging
Others
Clock
Averaging
Purging time: 0 min
Fuel setup
Purging
Unit setup
Passwords
Defaults
Enter
Enter
F8.20.ai
Figure 8.22
8.6.6
Other Settings
F8.21E.ai
Figure 8.23
Purging Display
Setting Passwords
Unauthorized access to lower level menu displays from the Execution/Setup display can be
protected by passwords. You can set separate passwords for Calibration, Blow back, and
Maintenance and for Commissioning
Proceed as follows:
(1) From the Basic panel display touch the [Setup] key, and the Execution/Setup display
appears.
(2) Select “Commissioning” , and the Commissioning display appears.
(3) Select “Others” then “Passwords”. The display shown in Fig. 8.22 appears.
First set the password for Calibration, Indication Check, Blow back, and Maintenance:
(4) Select “Calibration, Blow back, and Maintenance”.
(5) A text entry display appears. Enter password as up to 8 alphanumeric characters.
(6) You can enter a password for Commissioning by the same procedure.
(7) Record the passwords and look after them carefully.
Passwords
Calibration,Blow back
and Maintenance:
Commissioning:
Enter
F8.22E.ai
Figure 8.24
Passwords Display
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<8. Detailed Data Setting>
8-23
The passwords are not set as shipped from factory. If you reset initialize the parameters, and
password settings are deleted.
If you forget a password, select “Commissioning” in the Execution/Setup display, and enter
“MOON.” By doing so, you can enter the Commissioning display only. Then display the
Passwords and verify the set passwords.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
Blank Page
<9. Calibration>
9-1
9. Calibration
9.1
Calibration Briefs
9.1.1
Principle of Measurement with a zirconia oxygen
analyzer
This section sets forth the principles of measurement with a zirconia oxygen analyzer before
detailing calibration.
A solid electrolyte such as zirconia allows the conductivity of oxygen ions at high temperatures.
Therefore, when a zirconia-plated element with platinum electrodes on both sides is heated up
in contact with gases having different oxygen partial pressures on each side, the element shows
the action of the concentration cell. In other words, the electrode in contact with a gas with a
higher oxygen partial pressure acts as a negative electrode. As the gas comes in contact with the
zirconia element in this negative electrode, oxygen molecules in the gas acquire electrons and
become ions. Moving in the zirconia element, they eventually arrive at the positive electrode on
the opposite side.
There, the electrons are released and the ions return to the oxygen molecules. This reaction is
indicated as follows:
Negative electrode: O2 + 4e
Positive electrode: 2 O 2-
2 O 2O2 + 4 e
The electromotive force E (mV) between the two electrodes, generated by the reaction, is
governed by Nernst’s equation as follows:
E = -RT/nF ln Px/Pa........................................Equation (1)
where,
R: Gas constant
T: Absolute temperature
n: 4
F: Faraday’s constant
Px: Oxygen concentration in a gas in contact with the positive zirconia
electrode (%)
Pa: Oxygen concentration in a gas in contact with the negative zirconia
electrode (%)
Assuming the zirconia element is heated up to 750 °C, then we obtain equation (2) below.
E = −50.74 log Px ................................. Equation (2)
Pa
With this analyzer, the sensor (zirconia element) is heated up to 750°C, so Equation (2) is valid.
At that point, the relationship as in Figure 9.1 is effected between the oxygen concentration of the
sample gas in contact with the positive electrode and the electromotive force of the sensor (cell),
where a comparison gas of air is used on the negative electrode side.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
9-2
<9. Calibration>
120
100
0.51 vol% O2,81.92mV(Zero origin of calibration)
80
60
Cell
voltage
(mV)
40
20
21.0 vol% O2, 0mV
(Span origin of calibration)
0
-20
-40
0.1
0.5
1
5
10
21.0
50
100
Oxygen concentration (vol % O2)
Figure 9.1
F9.1E.EPS
Oxygen concentration in a Sample Gas vs Cell Voltage (21 vol%O2 Equivalent)
The measurement principles of a zirconia oxygen analyzer have been described above.
However, the relationship between oxygen concentration and the electromotive force of a cell is
only theoretical. Usually, in practice, a sensor shows a slight deviation from the theoretical value.
This is the reason why calibration is necessary. To meet this requirement, an analyzer calibration
is conducted so that a calibration curve is obtained, which corrects the deviation from the
theoretical cell electromotive force.
9.1.2
Measurement Principle of Zirconia Humidity Analyzer
A solid electrolyte such as zirconia allows the conduction of oxygen ions at high temperatures.
Therefore, when a zirconia-plated element with platinum electrodes on both sides is heated up
in contact with gases having different partial-oxygen pressures on each side, oxygen ions flow
from a high partial-oxygen pressure to a low partial-oxygen pressure, causing a voltage. When
a sample gas introduced into the zirconia-plated element with the measurement electrode, and
air (21.0 vol % O2) is flowed through the reference electrode, an electromotive force (mV) is
produced between the two electrodes, governed by Nernst’s equation as follows:
E = - RT/nF log e y/a ………………………… Equation (1)
where, R = Gas constant
T = Absolute temperature
n: 4
F = Faraday’s constant
y = O2 vol % on the zirconia element measurement electrode
a = O2 vol % to 21.0 vol % O2 on the zirconia element reference electrode
The humidity analyzer uses a sample gas composed of water vapor and air.
(A) For the vol % H2O measurement
x:Assuming that H2O vol % in a mixed gas is measured:
y = (100 – x) 3 0.21 …………………. Equation (2)
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
9-3
<9. Calibration>
From the above equations (1) and (2), we obtain:
E = -K log y/a = -Klog [(100 – x) 30.21] /21
= - K log (1 –0.01 x) ……………… Equation (3)
where, K = Constant
Using the above equation (3), we can calculate the water vapor in vol % from the electromotive
force.
Zirconia element
Comparison
water vapor
Sample gas
+ H 2O
-
Electrode
Air
100%
Water vapor
100%
H2O
x%
79%
100%
concentration
indicator
N2
O2
y%
21%
Sample gas composition
F9-1E.ai
Figure 9.2
Schematic Diagram of Measurement Principle
(B) For the “mixing ratio” measurement
Assuming that the mixing ratio is rkg/kg, then “r” can be calculated from the value of H2O vol %
as follows:
r = 0.622 3 x/(100 – x) …………… Equation (4)
From the above equations (1), (2) and (4), we obtain:
E = -K log y/a = -K log 50.622 3 21/(0.622 + r)/216
= -K log 0.622/(0.622 + r) … …… Equation (5)
where, K = Constant
With Equation (5), we can obtain the mixing ratio rkg/kg from the electromotive force.
120
Cell output, mV
E = -50.74 log PX/20.6
100
80
60
40
20
0
0.1
1
10
100
Oxygen concentration PX (%O2)
Oxygen concentration vs. cell output
F9-2E.ai
Figure 9.3
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<9. Calibration>
100
1.0
80
Mixing ratio, kg/kg
0.9
0.8
70
0.7
60
0.6
50
0.5
40
0.4
30
0.3
20
0.2
10
0.1
0
0
1
2
3
4
5
6
kg/kg
Humidity, vol % H2O
Mixing ratio
90
Humidity, vol % H2O
9-4
0.0
7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
Oxygen concentration, vol % O2
Oxygen concentration vs. Humidity, Mixing ratio
F9-3E.ai
Figure 9.4
9.1.3
Calibration Gas
A gas with a known oxygen concentration is used for calibration. Normal calibration is performed
using two different gases: a zero gas of low oxygen concentration and a span gas of high oxygen
concentration. In some cases, only one of the gases needs to be used for calibration. However,
even if only one of the gases is normally used, calibration using both gases should be done at
least once.
The zero gas normally used has an oxygen concentration of 0.95 to 1.0 vol%O2 with a balance of
nitrogen gas (N2). The span gas widely used is clean air (at a dew-point temperature below -20°C
and free of oily mist or dust, as in instrument air).
For best accuracy, as the span gas use oxygen whose concentration is near the top of the
measurement range, in a nitrogen mixture.
9.1.4
Compensation
The deviation of a measured value from the theoretical cell electromotive force is checked by the
method in Figure 9.5 or Figure 9.6.
Figure 9.5 shows a two-point calibration using two gases: zero and span. Cell electromotive
forces for a span gas with an oxygen concentration p1 and a zero gas with an oxygen
concentration p2 are measured while determining the calibration curve passing between these
two points. The oxygen concentration of the sample gas is determined from this calibration
curve. In addition, the calibration curve corrected by calibration is compared with the theoretical
calibration curve for determining the zero-point correction ratio represented by B/A x 100 (%) on
the basis of A, B and C shown in Figure 9.5 and a span correction ratio of C/A x 100 (%). If the
zero-point correction ratio exceeds the range of 100 ± 30% or the span correction ratio becomes
larger than 0 ± 18%, calibration of the sensor becomes impossible.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<9. Calibration>
81.92
9-5
Zero origin
ez
e2
Cell
electromotive
force, mV
Calibration curve
before correction
e1
es
B
A
Corrected calibration
curve (theoretical
calibration curve)
C
0
21.0
p1
Span gas concentration
Span origin
p2
0.51
Zero gas concentration
Oxygen concentration (vol%O2)
Zero-point correction ratio = (B/A) x 100 (%) Correctable range: 100 ± 30%
Span correction ratio = (C/A) x 100 (%)
Correctable range: 0 ± 18%
F9.2E.ai
Figure 9.5 Calculation of a Two-point Calibration Curve and Correction Ratios using Zero and
Span Gases
Figure 9.6 shows a one-point calibration using only a span gas. In this case, only the cell
electromotive force for a span gas with oxygen concentration p1 is measured. The cell
electromotive force for the zero gas is carried over from a previous measurement to obtain the
calibration curve. The principle of calibration using only a span gas also applies to the one-point
calibration method using a zero gas only.
81.92
Zero origin
ez
Cell
electromotive
force, mV
e1
es
Calibration curve before
correction
Previous
zero gas data
B
A
Corrected calibration curve
(theoretical calibration curve)
C
0
21.0
p1
Span gas concentration
0.51
Span origin
Oxygen concentration (vol%O2)
Zero point correction ratio = (B/A) x 100 (%) Correctable range: 100 ± 30%
Span correction ratio = (C/A) x 100 (%)
Correctable range: 0 ± 18%
F9.3E.ai
Figure 9.6
9.1.5
Calculation of a One-point Calibration Curve and Correction Ratios using a Span Gas
Characteristic Data from a Sensor Measured During
Calibration
During calibration, calibration data and sensor status data (listed below) are acquired.
However, if the calibration is not properly conducted (an error occurs in automatic or semiautomatic calibration), these data are not collected in the current calibration.
These data can be observed by selecting the [Detailed-data] key from the Basic panel display.
For an explanation and the operating procedures of individual data, consult Section 10.1,
Detailed-data Display.
(1) Record of span correction ratio
Recorded the past ten span correction ratios.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<9. Calibration>
9-6
(2) Record of zero correction ratio
Recorded the past ten zero correction ratios.
(3) Response time
You can monitor the response time provided that a two-point calibration has been done in semiautomatic or automatic calibration.
(4) Cell’s internal resistance
The cell’s internal resistance gradually increases as the cell (sensor) deteriorates. You can
monitor the values measured during the latest calibration. However, these values include the
cell’s internal resistance and other wiring connection resistance. So, the cell’s degrading cannot
be estimated from these values only.
When only a span calibration has been made, these values will not be measured, and previously
measured values will remain.
(5) Robustness of a cell
The robustness of a cell is an index for predicting the remaining life of a sensor and is expressed
in a number on four levels.
9.2
Calibration Procedures
NOTE
Calibration should be made under normal operating conditions (if the probe is connected to a
furnace, the analyzer will undergo calibration under the operating conditions of the furnace).
To make a precise calibration, conduct both zero and span calibrations.
The following sets forth the required calibration settings:
9.2.1 Mode
There are three calibration modes available:
(1) Manual calibration which allows zero and span calibrations or either one manually in turn;
(2) Semi-automatic calibration which lets calibration start with the touchpanel or a contact input,
and undergoes a series of calibration operations following preset calibration interval and
stabilization time.
(3) Automatic calibration which is carried out automatically following preset calibration interval.
Calibrations are limited by the following mode selection:
• When Manual calibration is selected:
Manual calibration only can be conducted. (This mode does not allow semi-automatic calibration
with a contact input nor automatic calibration even when its start-up time has reached.)
• When Semi-automatic calibration is selected:
This mode enables manual and semi-automatic calibrations to be conducted. (The mode,
however, does not allow automatic calibration even when its start-up time has reached.)
• When Automatic calibration is selected:
This calibration can be conducted in any mode.
To execute this calibration, follow these steps:
(1) Select the [Setup] key from the Basic panel display to display the Execution/Setup display.
Then select “Maintenance” from the Execution/Setup display.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
9-7
<9. Calibration>
(2) Select “Calibration setup” from the Maintenance display. Then select “Mode” from the
Calibration setup display (see Figure 9.7).
Now you can select “Manual”, “Semi_Auto”, or “Auto” calibration.
9.2.2
Calibration Procedure
Select both span and zero calibrations or span calibration only or zero calibration only.
Usually select span and zero calibrations.
Select “Points” from the Calibration setup display and then you can select “Both,” “Span” or
“Zero” (see Figure 9.8).
Calibration setup
Calibration setup
Mode: Manual
Points: Both
Span
Zero gas
conc:
1.00%
Zero
Span gas conc: 21 . 0 0 %
Timing
Mode: Manual
Points: Semi_Auto
Auto
Zero gas conc:
1.00%
Span gas conc: 2 1 . 0 0 %
Timing
Enter
Enter
F9.4E.ai
Figure 9.7
9.2.3
Calibration Setup display
F9.5E.ai
Figure 9.8
Calibration Setup display
Zero gas Concentration
Set the oxygen concentration for zero-point calibration. Enter the oxygen concentration for the
zero gas in the cylinder used in the following procedures:
Select “Zero gas conc” from the Calibration setup display. The numeric-data entry display then
appears. Enter the desired oxygen concentration for the zero-point calibration. (The zero gas set
ranges from 0.3 to 100 vol%O2.)
Enter 00098 for an oxygen concentration of 0.98 vol%O2 .
9.2.4
Span gas Concentration
Set the oxygen concentration for span calibration. If instrument air is used as the span gas, enter
21% O2.
Select “Span gas conc” from the Calibration setup display. Enter the desired span gas oxygen
concentration from the numeric-data entry display.
(The span gas set ranges from 4.5 to 100 vol%O2.)
Enter 02100 for an oxygen concentration of 21 vol%O2.
Instrument air is here defined as dry air with a dew-point temperature of no higher than -20°C. If
the dew-point temperature is higher than -20°C, use a hand-held oxygen analyzer to measure
the actual oxygen concentration.
When using the ZO21S Standard Gas Unit (for use of the atmospheric air as a span gas), use a
hand-held oxygen analyzer to measure the actual oxygen concentration, and then enter it.
NOTE
• When the instrument air supply is to be used as span gas, cool it to -20°C below dew point
to remove moisture, oil mist and dust from the air.
• If you do not do this to purify the air, then the accuracy of the calibration may be affected.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<9. Calibration>
9.2.5
9-8
Setting Calibration Time
• When the calibration mode is in manual:
First set the “Hold time” (output stabilization time). This indicates the time required from the
end of calibration to entering a measurement again. This time, after calibration, the sample
gas enters the sensor to set the time until the output returns to normal. The output remains
held after completing the calibration operation until the hold (output stabilization) time elapses.
The calibration time set ranges from 00 minutes, 00 seconds to 60 minutes, 59 seconds. For
more details, consult Section 8.2, Output Hold Setting. When the calibration mode is in semiautomatic, set the hold (output stabilization) time and calibration time. The calibration time is the
time required from starting the flow of the calibration gas to reading out the measured value. The
set calibration time is effective in conducting both zero and span calibrations. The calibrationtime set ranges from 00 minutes, 00 seconds to 60 minutes, 59 seconds. Figure 9.9 shows the
relationship between the calibration time and hold (output stabilization) time.
Calibration start
(contact or switch input)
Span calibration
(span gas valve open)
Calibration time
Zero calibration
(zero gas valve open)
Calibration time
Analog output status
Hold (output
stabilization) time
Analog output remains held
(when output remains held)
Figure9.9
F9.6E.ai
Calibration and Hold Time Settings
• When the calibration mode is in automatic:
In addition to the above Hold (output stabilization) time and Calibration time, set the Interval, Start
date, and Start time.
Interval means the calibration intervals ranging from 000 days, 00 hours to 255 days, 23 hours.
Set the first calibration day and the start-calibration time to the “Start date” and “Start time”
respectively. For example, to start the first calibration at 1:30 p.m. on March 25, 2001, enter
25/03/01 to the start date and 13 hours, 30 minutes to the start time, following the steps below:
(1) Select the “Calibration timing” display. A panel display as shown in Figure 9.10 appears.
(2) Select each item for the calibration to display the numeric-data entry display. Enter the
desired numeric values for the calibration.
Calibration timing
Hold time: 1 0 min 0 0 s
Cal time: 1 0 min 0 0 s
lnterval:
30d 00h
Start date: 0 1 / 0 1 / 0 0
Start time: 0 0 : 0 0
Enter
F9.7E.ai
Figure 9.10
Calibration Timing Display
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<9. Calibration>
9-9
NOTE
When setting calibration timing requirements, bear the following precautions in mind:
(1) If the calibration interval is shorter than the sum of hold (output stabilization) time plus
calibration time, the second calibration start time will conflict with the first calibration. In such
a case, the second calibration will not be conducted. (When both zero and span calibrations
are to be performed, the calibration time is double that required for a single (zero or span)
calibration.)
(2) For the same reason, if the calibration start time conflicts with manual calibration or semiautomatic calibration, the current calibration will not be conducted.
(3) If the calibration time conflicts with maintenance service or blow back operations, calibration
will start after completing the maintenance service or blow back operations (see Section
8.2.1, earlier in this manual).
(4) If 000 days, 00 hours are set for the calibration intervals, only the first calibration will be
conducted; a second or later calibration will not be conducted.
(5) If a past date is set to the calibration start day, no calibration will be conducted.
Default Values
When the analyzer is delivered, or if data are initialized, the calibration settings are by default, as
shown in Table 9.1.
Table 9.1
Default Settings for Calibration
Item
Default Setting
Calibration mode
Manual
Calibration procedure
Span - zero
Zero gas (oxygen) concentration
1.00%
Span gas (oxygen) concentration
21.00%
Hold (output stabilization) time
10 minutes, 00 seconds
Calibration time
10 minutes, 00 seconds
Calibration interval
30 days, 00 hours
Start date
01 / 01 / 00
Start time
0:00
9.3 Calibration
NOTE
(1) Perform calibration under normal working conditions (e.g. continuous operation with sensor
mounted on furnace).
(2) Perform both Span and Zero calibration for best resultant accuracy.
(3) When instrument air is used for the span calibration, remove the moisture from the
instrument air at a dew-point temperature of -20°C and also remove any oily mist and dust
from that air.
(4) If dehumidifying is not enough, or if foul air is used, the measurement accuracy will be
adversely affected.
9.3.1
Manual Calibration
For manual calibration, consult Section 7.12, Calibration, earlier in this manual.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
9.3.2
9-10
<9. Calibration>
Semi-automatic Calibration
To start calibration, follow these steps:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
Then select “Calibration” from the Execution/Setup display. The Calibration display shown in
Figure 9.11 appears.
(2) Select “Semi-auto calibration” to display the Semi-auto calibration display shown in Figure 9.12.
(3) Select “Start calibration”. The display shown in Figure 9.13 appears, and then start calibration.
Calibration
Semi-auto calibration
Check settings
Zero gas conc:
1.00%
Span gas conc: 2 1 . 0 0 %
Cal time: 1 0 min 0 0 s
Go to setup to change value.
Start calibration
Cancel calibration
Manual calibration
Semi-auto calibration
Enter
Enter
F9.8E.ai
Figure 9.11
Calibration Display
F9.9E.ai
Figure 9.12
Semi-auto Calibration Display
Semi-auto calibration
0.5min./div
2 1 . 0 0%
1 . 0 0%
2 1 . 0 % O2
CAL. TIME
F9.10E.ai
Figure 9.13
Semi-auto Calibration Display
To start calibration using an input contact, follow these steps:
(1) Make sure that Calibration start has been selected in the Input contacts display (see Section
8.5, earlier in this manual).
(2) Apply an input contact to start calibration.
To stop calibration midway, follow these steps:
(1) Press the [Reject] key. If this key is pressed midway during calibration, the calibration will
stop and the hold (output stabilization) time will be set up.
(2) Press the [Reject] key once again to return to the Basic panel display and the analyzer will
be in normal measurement.
9.3.3
Automatic Calibration
No execution operations are required for automatic calibration. Automatic calibration starts in
accordance with a preset start day and time. Calibration is then executed at preset intervals.
NOTE
Before starting Semi-automatic calibration or Automatic calibration, operate the calibration gas
solenoid valves and adjust calibration gas flow to 600 ± 60 ml/min.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<10. Other Functions>
10-1
10. Other Functions
10.1
Detailed-data Display
Press the [Detailed-data] key on the Basic panel display to view the detailed operation data as
shown in Figure 10.1.
Pressing the [▼] or [▲] key, you can advance the page or go back to your desired page.
• Detailed-data display
There are ten panel displays for viewing detailed data. The following briefly describe the
operational data displayed on the Detailed-data display.
Tag:
Span gas ratio:
0.0%
Zero gas ratio
100.0%
Response time:
0s
Cell robustness: life> 1year
Cell temperature:
750 °C
C.J.temperature:
47 °C
Hold
F10.1E.ai
Figure 10.1
10.1.1
Detailed-data Display
Span gas and Zero gas Correction Ratios
These are used to check for degradation of the sensor (cell). If the correction ratio is beyond the
limits as shown in Figure 10.2, the sensor should no longer be used.
These ratios can be found by calculating the data as shown below.
81.92
Zero origin
ez
Cell
electromotive
force, mV
Calibration curve
before correction
Previous
zero gas data
B
A
e1
Corrected calibration curve
(theoretical calibration curve)
es
C
0
21.0
p1
Span gas
concentration
0.51
Span origin
Oxygen concentration (vol%O2)
Zero gas correction ratio = (B/A) x 100 (%) Correctable range: 100 ± 30%
Span gas correction ratio = (C/A) x 100 (%) Correctable range: 0 ± 18%
F10.2E.ai
Figure 10.2
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<10. Other Functions>
10.1.2
10-2
Cell Response Time
The cell’s response time is obtained in the procedure shown in Figure 10.3. If only either a zeropoint or span calibration has been carried out, the response time will not be measured just as it
will not be measured in manual calibration.
Five minutes maximum
Response time
mA
90%
100%
10% of analog
output span
Time
Start calibration
Calibration
complete
The response time is obtained after the corrected calibration curve has been found. The response time
is calculated, starting at the point corresponding to 10% of the analog output up to the point at 90% of
the analog output span. That is, this response time is a 10 to 90% response.
F10.3E.ai
Figure 10.3
10.1.3
Typical Response Time characteristic
Robustness of a Cell
The robustness of a cell is an index for predicting the remaining life of a sensor and is expressed
as one of four time periods during which the cell may still be used:
(1) more than a year
(2) more than six months
(3) more than three months
(4) less than one month
The above four time periods are tentative and only used for preventive maintenance, not for
warranty of the performance.
This cell’s robustness can be found by a total evaluation of data involving the response time, the
cell’s internal resistance, and calibration factor. However, if a zero or span calibration was not
made, the response time cannot be measured. In such a case, the response time is not used as
a factor in evaluating the cell’s robustness.
10.1.4
Cell Temperature
This displays the cell (sensor) temperature, which is determined from the thermocouple emf and
cold junction temperature. Normally it is 750°C.
10.1.5
C. J. Temperature
This indicates the detector terminal box temperature, which compensates for the cold junction
temperature for a thermocouple measuring the cell temperature. When the ZR22 Detector is
used, the maximum C. J. temperature will be 150°C. If the terminal box temperature exceeds
this, take measures to shield the terminal box from heat radiation.
The maximum C. J. temperature varies depending on the type of detector.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<10. Other Functions>
10.1.6
10-3
Cell Voltage
The cell (sensor) voltage will be an index to determine the amount of degradation of the
sensor. The cell voltage corresponds to the oxygen concentration currently being measured.
If the indicated voltage approximates the ideal value (corresponding to the measured oxygen
concentration), the sensor will be assumed to be normal.
The ideal value of the cell voltage (E), when the oxygen concentration measurement temperature
is controlled at 750°C, may be expressed mathematically by:
E = -50.74 log (Px/Pa) [mV]
where, Px: Oxygen concentration in the sample gas
Pa: Oxygen concentration in the reference gas, (21 vol%O2)
Table 10.1 shows oxygen concentration versus cell voltage.
Table 10.1
Oxygen Concentration Vs. Cell Voltage, (cell temperature: 750 ºC)
%O2
mv
0.1
117.83
0.2
102.56
0.3
93.62
0.4
87.28
0.5
82.36
0.6
78.35
0.7
74.95
0.8
72.01
0.9
69.41
%O2
mv
1
67.09
2
51.82
3
42.88
4
36.54
5
31.62
6
27.61
7
24.21
8
21.27
9
18.67
21.0
0
30
-7.86
40
-14.2
50
-19.2
60
-23.1
70
-26.5
80
-29.5
90
-32.1
%O2
mv
%O2
mv
10.1.7
10
16.35
100
-34.4
T10.1E.ai
Thermocouple Voltage
The cell temperature is measured with a Type K (chromel-alumel) thermocouple. The
thermocouple cold junction is located in the detector terminal box. The cell temperature and the
thermocouple voltage (including the voltage corresponding to the cold junction temperature) are
displayed.
10.1.8
Cold Junction Resistance (C.J. Voltage)
The ZR22 Detector measures the cold junction temperature using an RTD (Pt 1000). (The earlier
model of ZO21D uses transistors to measure the cold junction temperature.)
If detector is “ZR22” is selected in the Basic setup display, the RTD resistance values will be
displayed. If the ZO21D is selected, the transistor voltage will be displayed.
10.1.9
Cell’s Internal Resistance
A new cell (sensor) indicates its internal resistance of 200 Ω maximum. As the cell degrades, so
will the cell’s internal resistance increase. The degradation of the cell cannot be evaluated just by
changes in cell’s internal resistance, however. Those changes in the cell’s internal resistance are
just a guide to the extent the cell is degrading. The updated values obtained during the calibration
are displayed.
10.1.10 Software Revision
The revision (number) of the software installed is displayed.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<10. Other Functions>
10-4
10.1.11 Maximum Oxygen Concentration, Humidity and
Mixing Ratio
The maximum oxygen concentration, humidity and mixing ratio and the time of its occurrence
during the period specified in the Averaging display are displayed. After the present monitoring
interval has elapsed, the maximum oxygen concentration that has been displayed so far will be
cleared and a new maximum oxygen concentration will be displayed. If the setup period of time
is changed, the current maximum oxygen concentration will be displayed (for more details, see
Section 8.6.2 earlier in this manual).
10.1.12 Minimum Oxygen Concentration, Humidity and Mixing
Ratio
The minimum oxygen concentration, humidity and mixing ratio and the time of its occurrence
during the period specified in the Averaging display are displayed. If the setup period elapses,
the minimum oxygen concentration that has been displayed so far will be cleared and a new
minimum oxygen concentration will be displayed. If the setup period of time is changed, the
current minimum oxygen concentration will be displayed (for more details, see Section 8.6.2
earlier in this manual).
10.1.13 Average Oxygen Concentration, Humidity and Mixing
Ratio
The average oxygen concentration, humidity and mixing ratio during the periods over which
average values are calculated is displayed. If the setup period elapses, the average oxygen
concentration that has been displayed so far will be cleared and a new average oxygen
concentration will be displayed. If the setup period of time is changed, the current average
oxygen concentration will be displayed (for more details, see Section 8.6.2 earlier in this manual).
10.1.14 Heater On-Time Ratio
The probe sensor is heated to and maintained at 750°C. When the sample gas temperature is
high, the amount of heater ON-time decreases.
10.1.15 Time
The current date and time are displayed. These are backed up with built-in batteries, so the clock
continue the run even if the power is switched off.
10.1.16 History of Calibration Time
The calibration-conducted dates and times, and span gas and zero gas ratios for the past ten
calibrations are stored in memory.
10.1.17 Power Supply Voltage
For the temperature control for the heater of the detector to work best, you should set the power
supply voltage and frequency appropriately, as the control parameters are based on this. Set the
AC supply voltage to “Low” if supply is 140 V AC or less, and to “High” if it is 180 V or more.
10.1.18 Power Frequency
Set the AC supply frequency setting appropriately --”Low” for 50 Hz, and “High” for 60 Hz.
10.2
Trend Graph
Press the [Graph display] key in the Basic panel display to switch to the graph display.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<10. Other Functions>
10-5
This will help grasp the measured-value trend. Touching anywhere on the graph display will
return to the Basic panel display. To set the Trend graph display, follow the steps.
10.2.1
Setting Display Items
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
Select “Maintenance” from the Execution/Setup display.
(2) Select “Display setup” from the Maintenance display.
(3) Select “Trend graph” from the Display setup display. The Trend graph display shown in
Figure 10.4 appears.
(4) Select “Parameter: Oxygen” from the Trend graph display. Then select the desired display
item shown in Table 10.2.
Table 10.2
Trend Graph Display Items
Selected item
Description
Oxygen concentration
Oxygen concentration graph under measurement
Humidity
Humidity graph under measurement
Mixing ratio
Mixing-ratio graph under measurement
Output 1
Output 1-selected item graph
If this equipment is for the oxygen analyzer, the trend graph will be an
oxygen concentration graph.
Output 2
Output 2-selected item graph
If this equipment is for the oxygen analyzer, the trend graph will be an
oxygen concentration graph.
Trend graph
Parameter:
Oxygen
Sample interval:
Scale Oxygen
Upper limit: 2 5 . 0
Lower limit:
0.0
30s
% O2
% O2
Enter
F10.4E.ai
Figure 10.4
10.2.2
Trend Graph Display
Sampling Interval
To plot a graph, set the “Sampling interval” for the measurement data.
This graph allows the plotting of 60 data items on one division on the time axis. So, if you set
a ten-second sampling interval, one division corresponds to 600 seconds (Figure 10.5). The
allowable sampling intervals range from 1 to 30 seconds. If you set a one-second sampling
interval, the axis of the abscissas then corresponds to five minutes.
If you set it to 30 seconds, the axis of the abscissas then corresponds to 150 minutes.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<10. Other Functions>
Preset upper limit
25.0 vol%O2
10 min./div
Preset lower limit
0.0 vol%O2
12.3 vol%O2
10-6
Time per division,
calculated by the set
sampling interval
Currently measured
value
60 data
Sampling period
Figure 10.5
10.2.3
F10.5E.ai
Plotting Graph for Sampling Interval
Setting Upper and Lower Limit Values on Graph
Set upper- and lower-limit values on the graph in the following procedure:
Press “Upper limit” in the Trend graph display. The numeric-data entry key appears.
Enter the upper-limit value. Also enter the lower-limit value in the same way. The allowable
settings for both upper-limit and lower-limit values range from 0 to 100 vol%O2 (0 to 100 %H2O
for humidity, 0 to 1 kg/kg for mixing ratio.)
10.2.4
Default Setting
When the analyzer is delivered, or if data are initialized, the set data are by default, as shown in
Table 10.3.
Table 10.3
Default Values for Graph Setting
Item Default
Value
Parameter
Oxygen concentration
Sampling interval
30 seconds
Upper limit (oxygen concentration)
25 vol%O2
Lower limit (oxygen concentration)
0 vol%O2
Upper limit (humidity)
25 % H2O
Lower limit (humidity)
0 % H2O
Upper limit (mixing ratio)
0.2 kg/kg
Lower limit (mixing ratio)
0 kg/kg
NOTE
If a rapid change in the measured value occurs during sampling, no sampled data are plotted on
the graph. Use the graph indication tentatively. Check the output current for accurate data.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
10.3
10.3.1
10-7
<10. Other Functions>
Other Display-related Functions
Auto-Return Time
On the Execution/Setup display or lower level menu displays, if no keys are touched for a preset
time, the Auto return time, then the display will automatically revert to the Basic panel display. The
“Auto return time” can be set in the range 0 to 255 minutes. If it is set to 0, then the display does
not automatically revert. By default, the “Auto return time” is set to 0 (zero).
(1) On the Basic panel display, touch the [Setup] key to display the Execution/Setup display,
then select “Maintenance”. Select “Display setup” then “Auto return time”.
(2) The numerical entry display appears for you to enter the desired Auto return time.
To set an Auto return time of one hour, enter 060.
(3) If you set 0, then the Auto return function does not operate.
The Auto return time is set to 0 prior factory shipment or after data initialization.
10.3.2
Entering Tag Name
You can attach a desired tag name to the equipment. To attach it, follow these steps:
(1) Select the [Setup] key from the Basic panel display to display the Execution/Setup display.
Then select “Maintenance” from the Execution/Setup display.
(2) Select the “Display setup” from the Maintenance display.
(3) Select the “Display item” from the Display setup display. The display shown in Figure 10.6
then appears.
(4) Select the “Tag name” from the Display item. The text-data entry display then appears.
(5) Enter up to 12 alphanumeric characters including codes for the desired tag name.
Display setup
Display item
Primary value:
Oxygen
Secondary value:
mA-ouput1
Tertiary value:
mA-output2
Tag name:
Display item
Trend graph
Auto return time:
0 min
Language: English
Deutsch
Francias
Enter
Enter
F10.6E.ai
Figure 10.6
10.3.3
Display Item Display
F10.7E.ai
Figure 10.7
Selecting the Display Language
Language Selection
You can select a display language from among English, Japanese, German and French.
The display language is set to the one specified in the purchase order when the analyzer is
shipped from the factory.
To select the language you want, follow these steps:
(1) On the Basic panel display, touch the [Setup] key to display the Execution/Setup display.
(2) Select “Maintenance”, then “Display setup”.
(3) Select “Language” and the dropdown selection (Figure 10.7) allows you to select the
desired language.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
10.4
<10. Other Functions>
10-8
Blow Back
This section explains the parameter settings for performing blow back.
10.4.1 Mode
There are three modes of blow back operation: no function, semi-automatic, and automatic.
Blow back is not performed when the mode is set to “No function”. In “Semi_Auto” mode, blow
back can be started by key operation on the display or by a contact input signal, and then
sequentially performed at a preset blow back time and hold time. In “Auto” mode, blow back
is automatically performed at preset intervals. For “Semi_Auto” or “Auto” modes, blow back is
performed. The following restrictions apply:
When “No function” is selected:
Blow back is not performed
When “Semi_Auto” is selected:
Semi-auto blow back can be performed. (Blow back does not start at Auto blow back start
time.)
When “Auto” is selected:
Blow back can be performed in either “Auto” or “Semi_Auto” mode.
(1) From the Basic panel display, touch [Setup] key; on the Execution/Setup display which
appears, select “Maintenance”.
(2) On the Maintenance display, select “Blow back setup” and the Mode selection pull down
allows you to select between “No function”, “Semi_Auto” and “Auto” (see Fig.10.8).
Blow back setup
Mode: No function
Semi_Auto
Auto
Enter
F10.9E.ai
Figure 10.8
10.4.2
Blow back setup Display
Operation of Blow back
Figure 10.9 shows a timing chart for the operation of blow back. To execute blow back with a
contact input, use a contact input with an ON-time period of one to 11 seconds.
Once blow back starts, a contact output repeatedly opens and closes at an interval of
approximately 10 seconds during the preset blow back time. After the blow back time elapses,
the analog output remains held at the preset status until the hold time elapses (refer to Section
8.2, earlier in this manual).
As the hold (output stabilization) time, set the time until the sample gas is returned to the sensor
and output returns to the normal operating conditions, after completing blow back operations.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<10. Other Functions>
10-9
ON-time period of 1 to 11 seconds for the contact input
Blow back start
(with contact input or
touchpanel operation)
Blow back time
ON
ON
(for 10
(for 10
seconds) OFF
seconds)
(for 10 seconds)
Contact output
(blow back switch)
ON
(for 10
seconds)
Hold time
Analog output remains held
(with HOLD feature specified)
Analog output
F10.9E.ai
Figure 10.9
10.4.3
Operation of Blow back
Setting Output Hold Time and Blow back Time
If the blow back mode is in “No function”, the output “Hold time” and “Blow back time” are not
displayed. If you select “Hold time”, the numeric-data entry display appears. Enter the desired
“Hold time” (output stabilization time) from 00 minutes, 00 seconds to 60 minutes, 59 seconds.
When you select “Blow back time”, the numeric-data entry display appears. Enter the desired
“Blow back time” from 00 minutes, 00 seconds to 60 minutes, 59 seconds.
10.4.4
Setting Interval, Start Date, and Start Time
The “Interval” is the time to execute blow back. Display the numeric-data entry panel display to
set the desired interval from 000 days, 00 hours to 255 days, 59 hours.
For the “Start date” and “Start time”, set the date when the blow back is first executed and the
time when to start the blow back, respectively. If you want to execute the first blow back, for
example, at 4:00 p.m. on March 25, 2001, enter 25/ 03/ 01 for the Start date and 16:00 for the
Start time.
Blow back setup
Mode:
Auto
Hold time: 1 0 min 0 0 s
Blow back time:
1 0 min 0 0 s
Interval:
30d 00h
Start date:
01/01/00
Start time:
00:00
Enter
F10.10E.ai
Figure 10.10
Blow back Setup Display
In the Blow back setup display shown in Figure 10.10, if you choose “Mode: No function” or
“Semi_Auto”, the Interval, Start date, and Start time for these are not displayed.
NOTE
• If the blow back is executed with an input contact, it must be preset in the Input contact
settings (for more details, see Section 8.5, earlier in this manual).
• In Section 8.4, Output Contact Setup, earlier in this manual, set the contact used as the
blow back switch beforehand.
• Do not set any other function for the contact used as the blow back switch.
Otherwise, blow back may be activated when the contact is closed by any other function.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
10-10
<10. Other Functions>
• No blow back is executed during calibration or maintenance service. If automatic blow back
reaches the preset start time during calibration or maintenance service, blow back will be
executed after completing the calibration or maintenance service and after the equipment
returns to the measurement mode.
• If automatic blow back reaches the preset start time during semi-automatic blow back, the
current automatic blow back will not be executed.
• If you set the blow back interval at 000 days, 00 hours, only the first blow back is then
executed. No subsequent blow backs will be executed.
• If a past date is set for the Start time, no blow back will be executed.
10.4.5
Default Setting
When the analyzer is delivered, or if data are initialized, the blow back settings are by default, as
shown in Table 10.4.
Table 10.4
Blow back Default Setting
Item
Default setting
Mode
No function (“invalid”)
Hold time
10 minutes, 00 seconds
Blow back time
10 minutes, 00 seconds
Interval
30 days, 00 hours
Start date
01/ 01/00
Start time
0:00
10.5
Parameter Initialization
Parameter settings can be initialized to the factory default settings. Initialization can be done
for all parameters or for individual parameters. The parameters that can be initialized and their
defaults are listed in Table 10.5.
(1) On the Basic panel display, touch the [Setup] key to display the Execution/Setup display.
(2) Select “Commissioning”, next “Others” then “Defaults”. A display like Figure 10.11. appears.
(3) Select the desired item to be initialized then a display like Figure 10.12 appears.
Select “Defaults start” then initialization starts.
Defaults
Display
Calibration
Blow back
mA-outputs
Alarms
Contacts
Others
All data
Defaults
Defaults start
Cancel
Enter
F10.11E.ai
Figure 10.11
Display
Defaults Display
Enter
Figure 10.12
Initialization Start Display
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<10. Other Functions>
10-11
WARNING
Do not turn off the power during initialization. Otherwise, initialization will not be performed
properly.
Table 10.5
Initialization Items and Default Values (Oxygen Analyzer)
Item
Equipment
selection
Initialization Parameter
Type of equipment
Not initialized
Detector
ZR22
Sample gas
Wet gas
Display item
Displayed data
Trend graph
Calibration
data
Blow back
Default setting
1st display item
Oxygen concentration
2nd display item
Current output 1
3rd display item
Current output 2
Tag name
Deleted
Parameter
Oxygen concentration
Sampling interval
30 seconds
Upper limit (graph)
25 vol%O2
Lower limit (graph)
0 vol%O2
Automatic return time
0 min.
Language
Not initialized
Calibration
setting
Blow back
setting
Mode
Manual
Calibration procedure
Span - zero
Zero gas concentration
1.00 vol%O2
Span gas concentration
21.00 vol%O2
Output hold time
10 min., 00 sec.
(Semi_Auto)
Calibration time
10 min., 00 sec.
(Semi_Auto)
Interval
30 days, 00 hr. (Auto)
Start date
01/01/00 (Auto)
Start time
0:00 (Auto)
Mode
No function (invalid)
(Output) hold time
10 min., 00 sec.
Blow back time
10 min., 00 sec.
Interval
30 days, 00 hr.
Start date
01/01/00
Start time
0:00
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<10. Other Functions>
Item
Initialization Parameter
Parameter
mA-output 1
mA-output 2
Current output
data
Output hold
setting
Alarm setting
Alarm data
Alarm set value
10-12
Default setting
Oxygen concentration
Min. oxygen concentration
0 vol%O2
Max. oxygen concentration
25 vol%O2
Output damping constant
0 sec.
Output mode
Linear
Warm-up
4 mA
Set value of Warm-up
4 mA
Maintenance
Previous value held
Set value of Maintenance
4 mA
Calibration, blow back
Previous value held
Set value of Calibration, blow back
4 mA
Error
Preset value held
Set value of Error
3.4 mA
Parameter
Oxygen concentration
Hysteresis
0.1 vol%O2
Delayed action of alarm contact
3 seconds
High-high alarm
OFF
Alarm value of High-high alarm
100 vol%O2
High limit alarm
OFF
Alarm value of High limit alarm
100 vol%O2
Low limit alarm
OFF
Alarm value of Low limit alarm
0 vol%O2
Low-low alarm
OFF
Alarm value of Low-low alarm
0 vol%O2
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<10. Other Functions>
Item
Initialization Parameter
Default setting
Alarm
Output contact 1
Output contact 2
Other
settings
OFF
Warm-up
ON
Output range now being switched
OFF
Now calibrating
OFF
Now maintenance servicing
ON
Blow back
OFF
High limit temp. alarm
OFF
Calibration gas press. drop
OFF
Gas leak detection
OFF
Contact output action
Open
Alarm
OFF
Other
settings
Contacts
Warm-up
OFF
Output range switching
OFF
Now calibrating
ON
Now maintenance servicing
OFF
Blow back
OFF
High limit temp. alarm
OFF
Calibration gas press. drop
OFF
Gas leak detection
OFF
Contact output action
Alarm
Output contact 3
Input contact 1
Input contact 2
Fuel set value
Password
High-high alarm
OFF
High limit alarm
ON
Low limit alarm
ON
Low-low alarm
OFF
Calibration Coefficient Alarm
OFF
Startup power stabilization timeout
OFF
Error
OFF
OFF
Contact output
Closed
Error
ON
Other settings
OFF
Contact output
Closed (Fixed)
Function
OFF
Action
Closed
Average value/Maximum and
minimum values
Other data
Closed
Other settings
Alarm
Output contact 4
10-13
Average-value calculation
One hour
Intervals over which max. and min. 24 hours
values are monitored
Amount of water vapor in exhaust
gas
0.00 m3/kg (m3)
Theoretical amount of air
1.00 m3/kg (m3)
X value
1.00
Absolute humidity of atmosphere
0.0 kg/kg
Deleted
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<10. Other Functions>
Table 10.6
Initialization Items and Default Values (Humidity Analyzer)
Item
Equipment
selection
Initialization Parameter
Not initialized
Detector
ZR22
Sample gas
Displayed data
Trend graph
Blow back
Default setting
Type of equipment
Display item
Calibration
data
10-14
Wet gas
1st display item
Humidity
2nd display item
Current output 1
3rd display item
Current output 2
Tag name
Deleted
Parameter
Humidity
Sampling interval
30 seconds
Upper limit (graph)
25 vol%O2
Lower limit (graph)
0 vol%O2
Automatic return time
0 min.
Language
Not initialized
Calibration
setting
Blow back
setting
mA-output 1
mA-output 2
Current output
data
Output hold
setting
Mode
Manual
Calibration procedure
Span - zero
Zero gas concentration
1.00 vol%O2
Span gas concentration
21.00 vol%O2
Output hold time
10 min., 00 sec.
(Semi_Auto)
Calibration time
10 min., 00 sec.
(Semi_Auto)
Interval
30 days, 00 hr. (Auto)
Start date
01/01/00 (Auto)
Start time
0:00 (Auto)
Mode
No function (invalid)
(Output) hold time
10 min., 00 sec.
Blow back time
10 min., 00 sec.
Interval
30 days, 00 hr.
Start date
01/01/00
Start time
0:00
Parameter
Humidity
Min. oxygen concentration
0 vol%O2
Max. oxygen concentration
25 vol%O2
Output damping constant
0 sec.
Output mode
Linear
Warm-up
4 mA
Set value of Warm-up
4 mA
Maintenance
Previous value held
Set value of Maintenance
4 mA
Calibration, blow back
Previous value held
Set value of Calibration, blow back
4 mA
Error
Preset value held
Set value of Error
3.4 mA
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<10. Other Functions>
Item
Initialization Parameter
Parameter
Alarm setting
Alarm set value
Humidity alarm
setting
Alarm data
Humidity alarm
set value
Mixing-ratio
alarm setting
Mixing-ratio
alarm set value
10-15
Default setting
Oxygen concentration
Hysteresis
0.1 vol%O2
Delayed action of alarm contact
3 seconds
High-high alarm
OFF
Alarm value of High-high alarm
100 vol%O2
High limit alarm
OFF
Alarm value of High limit alarm
100 vol%O2
Low limit alarm
OFF
Alarm value of Low limit alarm
0 vol%O2
Low-low alarm
OFF
Alarm value of Low-low alarm
0 vol%O2
Hysteresis
0.1 % H2O
Contact delay
3 seconds
High-high alarm
OFF
Alarm value of High-high alarm
100 % H2O
High limit alarm
OFF
Alarm value of High limit alarm
100 % H2O
Low limit alarm
OFF
Alarm value of Low limit alarm
0 % H2O
Low-low alarm
OFF
Alarm value of Low-low alarm
0 % H2O
Hysteresis
0 % kg/kg
Contact delay
3 seconds
High-high alarm
OFF
Alarm value of High-high alarm
1 kg/kg
High limit alarm
OFF
Alarm value of High limit alarm
1 kg/kg
Low limit alarm
OFF
Alarm value of Low limit alarm
0 kg/kg
Low-low alarm
OFF
Alarm value of Low-low alarm
0 kg/kg
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<10. Other Functions>
Item
Initialization Parameter
Default setting
Alarm
Output contact 1
Contacts
Output contact 2
Other
settings
OFF
Warm-up
ON
Output range now being switched
OFF
Now calibrating
OFF
Now maintenance servicing
ON
Blow back
OFF
High limit temp. alarm
OFF
Calibration gas press. drop
OFF
Gas leak detection
OFF
Contact output action
Open
Alarm
OFF
Other
settings
Warm-up
OFF
Output range switching
OFF
Now calibrating
ON
Now maintenance servicing
OFF
Blow back
OFF
High limit temp. alarm
OFF
Calibration gas press. drop
OFF
Gas leak detection
OFF
Contact output action
Alarm
Output contact 3
Input contact 1
Input contact 2
High-high alarm
OFF
High limit alarm
ON
Low limit alarm
ON
Low-low alarm
OFF
Calibration Coefficient Alarm
OFF
Startup power stabilization timeout
OFF
Error
OFF
OFF
Contact output
Closed
Function
OFF
Action
Closed
Other data
Measurement gas temperature
Password
Closed
Other settings
Average value/Maximum and
minimum values
10-16
Average-value calculation
One hour
Intervals over which max. and min. 24 hours
values are monitored
Temperature input mode
Preset
Measurement gas temperature
300°C
Temperature at 4 mA
0°C
Temperature at 20 mA
1000°C
Measurement gas pressure
101.33 kPa abs.
Deleted
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<10. Other Functions>
10.6
10-17
Reset
Resetting enables the equipment to restart. If the equipment is reset, the power is turned off
and then back on. In practical use, the power remains on, and the equipment is restarted under
program control. Resetting will be possible in the following conditions:
(1) Error 1 – if the cell voltage is defective
(2) Error 2 – if a temperature alarm occurs
(3) Error 3 – if the A/D converter is defective
(4) Error 4 – if an EEPROM write error occurs
For details on error occurrence, consult Chapter 12, Troubleshooting, later in this manual.
If any of the above problems occurs, the equipment turns off the power to the detector heater.
To cancel the error, reset the equipment following the steps below, or turn the power off and then
back on.
NOTE
Make sure that before resetting or restarting the power that there is no problem with the detector
or converter.
If a problem arises again after the resetting, turn the power off and troubleshoot the problem by
consulting the Troubleshooting chapter later in this manual.
To reset the equipment, follow these steps:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
(2) Choose “Reset”. The Reset display shown in Figure 10.13 appears.
(3) Choose “Start reset” and then press the [Enter] key to reset the equipment and the
equipment will then be in its warm-up state.
Reset
Start reset
Cancel reset
Enter
F10.13E.ai
Figure 10.13
10.7
Reset Display
Handling of the ZO21S Standard Gas Unit
The following describe how to flow zero and span gases using the ZO21S Standard Gas
Unit. Operate the ZO21S Standard Gas Unit, for calibrating a system classified as System 1,
according to the procedures that follow.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<10. Other Functions>
10.7.1
10-18
Standard Gas Unit Component Identification
Carrying case
Flow checker
Checks the zero and
span gas flow.
Span gas valve
Controls the span gas (air) flow.
Zero gas valve regulator
Cover screws (six pcs.)
Tube connection
Pump
Gas cylinder
Supplies span gas (air)
Contains the zero gas. A gas
of 7 Nl is charged to 700 kPa
Zero gas valve
Clamp
Attaches to the gas
cylinder for use.
Clamps the gas cylinder.
Power cord
Applies the power to operate the
pump to supply the span gas.
Figure 10.14
F10.14E.ai
Standard Gas Unit Component Identification
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
10.7.2
<10. Other Functions>
10-19
Installing Gas Cylinders
Each ZO21S Standard Gas Unit comes with six zero gas cylinders including a spare. Each
gas cylinder contains 7-liters of gas with a 0.95 to 1.0 vol%O2 (concentration varies with each
cylinder) and nitrogen, at a pressure of 700 kPaG (at 35°C).
The operating details and handling precautions are also printed on the product. Please read them
beforehand.
To install the gas cylinder, follow these steps:
(1) Attach the zero gas valves onto the gas cylinder. First, turn the valve regulator of the zero
gas valves counterclockwise to completely retract the needle at the top from the gasket
surface. Maintaining the valve in this position, screw the valve mounting into the mouthpiece
of the gas cylinder. (If screw connection is proper, you can turn the screw manually. Do not
use any tool.) When the gasket comes in contact with the mouthpiece of the gas cylinder
and you can no longer turn it manually, tighten the lock nut with a wrench.
(2) Remove the carrying case from the standard gas unit. The case is attached to the unit with
six screws. So, loosen the screws and lift them off.
(3) Slide the gas cylinder through the hole in the back of the unit and connect the tube (the
piping in the unit) to the valve connections. Insert each tube at least 10 mm to prevent
leakage, and secure it using a tube clamp.
(4) Attach the gas cylinder to the case. Extend the valve regulator of the zero gas valves
through the hole in the front panel of the unit and secure the bottom of the cylinder with the
clamp.
(5) Take note of the oxygen concentration of the sealed gas indicated on the gas cylinder and
replace the carrying case. Enter the oxygen concentration of the sealed gas, following
instructions on the converter display. Also check that no piping is disconnected.
Thus, the work of installing a gas cylinder is completed. However, gases in the cylinders cannot
immediately flow out after these procedures. To discharge the gases, it is necessary for the
needle in the zero gas valves to puncture a hole in the gas cylinder (see Section 10.7.3).
10.7.3
Calibration Gas Flow
(1) To operate the standard gas unit, place it on a nearly horizontal surface in order to allow the
flow check to indicate the precise flow rate. In addition, a power supply for driving the span
gas (air) supply pump is required near the unit (the length of the power cord attached to the
unit is 2 m). Select a suitable location for the unit near the installation site of the converter.
(2) Connect the tube connector port of the standard gas unit to the calibration gas inlet of the
detector, using a polyethylene resin tube with an outside diameter of 6 mm. Be careful to
prevent gas leakage.
(3) Fully open the stop valve mounted on the calibration gas inlet of the detector.
(4) Enter the oxygen concentration of the sealed gas (noted from the cylinder) into the
converter. Also check that the oxygen concentration of the span gas is correctly set
(21 vol%O2 for clean air). When using the ZO21S Standard Gas Unit (for use of the
atmospheric air as a span gas), use a hand-held oxygen analyzer to measure the actual
oxygen concentration, and then enter it.
The standard gas unit is used only when manual calibration is employed. Therefore, the timing
for flowing span gas (air) is included in the manual calibration flowchart described in Section
10.7.2. For operation of the converter, see Section 7.12, earlier in this manual.
(1) When the message “Open span gas valve” is displayed on the converter display during
calibration, plug the power cord into the power supply socket to start the pump of the
standard gas unit.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<10. Other Functions>
10-20
Manual calibration
Open span gas valve.
Set flow span gas to
600ml/min.
Valve opened
Cancel calibration
Enter
F10.15E.ai
Figure 10.15
Manual Calibration Display
(2) Next, adjust the flow rate to 600 ± 60 ml/min using the span gas valve “AIR” (the flow check
ball stops floating on the green line when the valve is slowly opened). To rotate the valve
shaft, loosen the lock nut and turn it using a flat-blade screwdriver. Turning the valve shaft
counterclockwise increases the flow rate.
(3) After adjusting the flow rate, tighten the valve lock nut.
(4) Select “Valve opened” (to start calibration) from the Manual calibration display shown in
Figure 10.15. Check the Trend graph display to see that the measured value is stabilized.
Then press the [Enter] key. The Manual calibration display shown in Figure 10.16 appears.
Disconnect the power cord to stop the pump.
Manual calibration
Span calibration
Close the span gas valve.
Zero calibration
End
Enter
F10.16E.ai
Figure 10.16
Manual Calibration Display
Cause a zero gas to flow according to the Manual calibration display shown in Figure 10.17.
Manual calibration
Open zero gas valve.
Set flow zero gas to
600ml/min.
Valve opened
Cancel calibration
Enter
F10.17E.ai
Figure 10.17
Manual Calibration Display
(1) Use the needle of the zero gas valve “ CHECK GAS “ to puncture a hole in the gas cylinder
installed as described in Section 10.7.2. Fully clockwise turn the valve regulator by hand.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<10. Other Functions>
10-21
(2) Next, adjust the flow rate to 600 ± 60 ml/min (the flow check ball stops floating on the green
line when the valve is slowly opened). Turn the regulator of the zero gas valves back slowly
counterclockwise. At that time, the flow rate also decreases as the inner pressure of the gas
cylinder decreases. Therefore, monitor the flow check and, when the ball’s position changes
greatly, readjust the valve.
(3) Select “Valve opened” (to start calibration) from the Manual calibration display. Check the
Trend graph display to see that the measured value is stabilized. Then press the [Enter] key.
The Manual calibration display shown in Figure 10.18 appears. Then stop the zero gas flow
immediately. Turn the zero gas valve regulator fully clockwise. If this valve regulator is not
properly adjusted, the needle valve will not close completely and a cylinder gas may leak.
Manual calibration
Zero calibration
Close the zero gas valve.
Span calibration
End
Enter
F10.18E.ai
Figure 10.18
Zero gas Calibration Complete (in Manual Calibration Display)
NOTE
Be sure not to terminate a calibration in progress because of a shortage of gas in the cylinder.
Each gas cylinder is operable for nine minutes or more provided the gas is discharged at the
specified rate. Therefore, if your calibration time is estimated at four minutes, you can operate the
zero-point calibration twice.
(1) Fully close the stop valve mounted on the calibration gas inlet of the detector.
(2) Remove the tube connecting the detector to the standard gas unit.
WARNING
Store the standard gas unit with the gas cylinder mounted where the ambient temperature does
not exceed 40°C. Otherwise, the gas cylinder may explode. Store the spare gas cylinders under
the same condition.
10.8
Methods of Operating Valves in the ZA8F
Flow Setting Unit
The ZA8F Flow Setting Unit is used as a calibration device for a system conforming to System
2. Calibration in such a system is to be manually operated. So, you have to operate the valve of
the Flow Setting each time calibration is made (starting and stopping the calibration gas flow and
adjusting the flow rate). This applies even if you are using the ZR40H Automatic Calibration Unit.
For operation of the converter, see Section 7.12, earlier in this manual.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<10. Other Functions>
10.8.1
10-22
Preparation Before Calibration
To operate the ZA8F Flow Setting Unit, prepare for calibration as follows:
(1) Check for a complete closing of the zero gas flow setting valve in the unit and open the
regulator valve for the zero gas cylinder until the secondary pressure equals sample gas
pressure plus approx. 50 kPa (or sample gas pressure plus approx. 150 kPa when a check
valve is used, maximum pressure rating is 300 kPa).
(2) Check that the oxygen concentration of the zero gas and span gas (instrument air 21
vol%O2) in the cylinder is set in the converter.
10.8.2
Operating the Span Gas Flow Setting Valve
The following description is given assuming that instrument air, the same as the reference gas, is
used as the span gas.
(1) When the display shown in Figure 10.15 appears during calibration, open the span gas
flow setting valve of the flow setting unit and adjust the flow rate to 600 ± 60 ml/min. Turn
the valve slowly counterclockwise after loosening the lock nut if the valve has a lock nut. To
check the flow rate, use the calibration flowmeter. If the sample gas pressure is extremely
high, adjust the sample gas pressure to obtain pressures (listed in Table 10.7) ± 10%.
Table 10.7
Sample gas pressure (kPa)
50
100
150
200
250
Flow rate (ml/min)
500
430
380
350
320
(2) Adjust the flow rate and select “Valve opened” from the Manual calibration display.
Check the Trend graph display to see that the measured value is stabilized. Then press the
[Enter] key. The Manual calibration display shown in Figure 10.16 appears.
(3) Close the span gas flow setting valve to stop the span gas (air) flow. If the valve has a lock
nut, be sure to tighten the lock nut to prevent any leakage of span gas into the sensor during
measurement.
10.8.3
Operating the Zero Gas Flow Setting Valve
Operate the zero gas flow setting valve during zero-point calibration in the following procedures:
(1) When the display shown in Figure 10.19 appears during calibration, open the zero gas
flow setting valve of the flow setting unit and adjust the flow rate to 600 ± 60 ml/ min. To
rotate the valve shaft, if the valve has a lock nut loosen the lock nut and slowly turn it
counterclockwise. To check the flow rate, monitor the calibration gas flowmeter.
If the sample gas pressure is extremely high, adjust the sample gas pressure to obtain
pressures (listed in Table 10.7) ± 10%.
Manual calibration
Open zero gas valve.
Set flow zero gas to
600ml/min.
Valve opened
Cancel calibration
Enter
F10.19E.ai
Figure 10.19
Manual Calibration Display
(2) Adjust the flow rate and select “Valve opened” from the Manual calibration display.
Check the Trend graph display to see that the measured value is stabilized. Then press the
[Enter] key. The Manual calibration display shown in Figure 10.20 appears.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<10. Other Functions>
10-23
Manual calibration
Zero calibration
Close the zero gas valve.
Span calibration
End
Enter
F10.20E.ai
Figure 10.20
Zero-point Calibration Complete (in Manual Calibration)
(3) Close the zero gas flow setting valve to stop the zero gas flow. If the valve has a lock nut,
be sure to tighten the lock nut to prevent the any leakage of the zero gas into the sensor
because the valve may become loose during measurement.
10.8.4
Operation After Calibration
No special operation of the instrument is needed after calibration. However, it is recommended
that the pressure reducing valve for the zero gas cylinders be closed because calibration is not
required so often.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
Blank Page
<11. Inspection and Maintenance>
11-1
11. Inspection and Maintenance
This chapter describes the inspection and maintenance procedures for the EXAxt ZR Zirconia
Oxygen Analyzer to maintain its measuring performance and normal operating conditions.
WARNING
Do not touch the probe if it has been in operation immediately just before being checked. (The
sensor at the tip of the probe heats up to 750°C during operation. If you touch it, you will get
burned.)
CAUTION
When checking the detector, carefully observe the following:
• Do not subject the probe to shock or cool it rapidly.
The sensor is made of ceramic (zirconia). If the detector is dropped or bumped into
something, the sensor may be damaged and no longer work.
• Do not reuse a metal O-ring to seal the cell assembly. If you replace the cell or remove it
from the probe for checking, be sure to replace the metal O-ring. Otherwise, the furnace gas
may leak, and then the leaking corrosive gas will cause the built-in heater or thermocouple
to go open circuit, or the detector may corrode.
• Handle the probe with care so that the dust-filter mounted screws on the tip of the probe do
not hurt your finger(s).
• Before opening or closing the terminal box, first remove dust, sand, or the like from the
terminal box cover.
11.1
11.1.1
Inspection and Maintenance of the Detector
Cleaning the Calibration Gas Tube
The calibration gas, supplied through the calibration gas inlet of the terminal box into the detector,
flows through the tube and comes out at the tip of the probe. The tube might become clogged
with dust from the sample gas. If you become aware of clogging, such as when a higher pressure
is required to achieve a specified flow rate (600 ± 60 ml/min), clean the calibration gas tube.
To clean the tube, follow these steps:
(1) Remove the detector from the installation assembly.
(2) Following Section 11.1.2, later in this manual, remove the four bolts (and associated
washers) that tighten the sensor assembly, and the pipe support as well as the U-shaped
pipe.
(3) Use a rod 2 to 2.5 mm in diameter to clean the calibration gas tube inside the probe. In
doing this, keep air flowing from the calibration gas inlet at about 600 ml/min and insert the
rod into the tube (3-mm inside diameter). However, be careful not to insert the rod deeper
than 40 cm for a general-purpose detector, or 15 cm for high temperature detector.
(4) Clean the U-shaped pipe. The pipe can be rinsed with water. However, it should be dried out
thoroughly before reassembly.
(5) Restore all components you removed for cleaning. Follow Section 11.1.2 to restore all
components in their original positions. Be sure to replace the O-ring(s) with new ones.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<11. Inspection and Maintenance>
Exploded view of components
Rod
(with outside diameter
of 2 to 2.5 mm)
11-2
Calibration gas tube
F11.1E.ai
Figure 11.1 Cleaning the Calibration Gas Tube
11.1.2
Replacing the Sensor Assembly
The performance of the sensor (cell) deteriorates as its surface becomes soiled during operation.
Therefore, you have to replace the sensor when its life expectancy expires, for example, when
it can no longer satisfy a zero gas ratio of 100 ± 30% or a span gas ratio of 0 ± 18%. In addition,
the sensor assembly is to be replaced if it becomes damaged and can no longer operate during
measurement.
If the sensor becomes no longer operable (for example, due to breakage), investigate the cause
and remedy the problem as much as possible to prevent recurrence.
CAUTION
• If the sensor assembly is to be replaced, allow enough time for the detector to cool down
from its high temperature. Otherwise, you may get burned.
• If the cell assembly is to be replaced, be sure to replace the metal O-ring and the contact
together.
Additionally, even in a case where the cell is not replaced, if the contact becomes deformed
and cannot make complete contact with the cell, replace the contact.
• If there is any corroded or discolored area in the metal O-ring groove in which the contact
is embedded, sand the groove with sandpaper or use a metal brush, and then sand further
with a higher grade of sandpaper (no. 1500 or so), or use an appropriate metal brush
to eliminate any sharp protrusions on the groove. The contact’s resistance should be
minimized.
• Use sensor assemblies manufactured in or after Sept. 2000: the serial number on the side
of the sensor assembly should be 0J000 or later (for example: 0K123, 1AA01 etc)
1.
Identifying parts to be replaced
In order not to lose or damage disassembled parts, identify the parts to be replaced from among
all the parts in the sensor assembly. Normally, replace the sensor, metal O-ring and contact
together at the same time. If required, also replace the U-shaped pipe, bolts, filter and associated
spring washers.
2.
Removal procedures
(1) Remove the four bolts and associated washers from the tip of the detector probe.
(2) Remove the U-shaped pipe support together with the U-shaped pipe. Remove filter also.
(3) Pull the sensor assembly toward you while turning it clockwise. Also, remove the metal
O-ring between the assembly and the probe. Remove filter also.
(When replacing the assembly, be careful not to scratch or dent the tip of the probe with
which the metal O-ring comes in contact (the surface with which the sensor flange also
comes in contact). Otherwise, the sample gas will not be sealed.)
(4) Use tweezers to pull the contact out of the groove in the tip of the probe.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<11. Inspection and Maintenance>
11-3
(5) Clean the sensor assembly, especially the metal O-ring contact surface to remove any
contaminants adhering to that part. If you can use any of the parts from among those
removed, also clean them up to remove any contaminants adhering to them.
(Once the metal O-ring has been used, it can not be reused. So, be sure to replace it.)
3.
Part assembly procedure
(1) First, install the contact. Being careful not to cause irregularities in the pitch of the coil spirals
(i.e., not to bend the coil out of shape), place it in the ringed groove properly so that it forms
a solid contact.
Groove in which the contact
(E7042BS) is placed
F11.2E.ai
Figure 11.2
Installing the Contact
(2) Next, make sure that the O-ring groove on the flange surface of the sensor is clean.
Install the metal O-ring in that O-ring groove, and then insert the sensor in the probe while
turning it clockwise. After inserting it until the metal O-ring comes in contact with the probe’s
O-ring contact surface, properly align the U-shaped pipe insertion holes with the bolt
openings.
(3) Attach the U-shaped pipe to its support with filter, then fully insert the U-shaped pipe and its
support into the probe.
(4) Coat the threads of the four bolts with anti-seize grease and then screw them in along with
the washers. First, tighten the four bolts uniformly by hand, and then use a torque wrench
to tighten all areas of the metal O-ring uniformly, that is, to make sure the sensor flange is
perfectly horizontal to the O-ring’s working face in the probe.
This is done by tightening first one bolt and then its opposing bolt each 1/8 turn, and then
one of the other bolts followed by its opposing bolt, each also 1/8 turn. This continues
in rotating fashion until they are all fully tightened with the torque wrench preset to
approximately 5.9 N • m. If they are not uniformly tightened, the sensor or heater may be
damaged.
Replacement of the sensor assembly is now complete. Install the detector and restart
operation. Calibrate the instrument before making a measurement.
IM 11M12A01-02E
12th
10thEdition
Edition::Apr.
Aug.14,2015-00
27, 2017-00
11-4
<11. Inspection and Maintenance>
Metal O-ring
Sensor
Dust filter
(optional)
Bolts (four)
U-shaped pipe
support
Contact
Probe
Filter
U-shaped pipe
Washers
(four)
1/8 turn – tighten bolts 1/8 turn
(approximately 45°) each
Figure 11.3
F11.3E.ai
Exploded View of Sensor Assembly
NOTE
Optional Inconel bolts have a high coefficient of expansion. If excess torque is applied while the
bolts are being tightened, abnormal strain or bolt breakage may result.
So, tighten the bolts following the instructions given above.
11.1.3
Replacement of the Heater Unit
This section describes the replacement procedure for the heater unit.
The sensor or ceramic heater-furnace core internal structure is subject to fracturing, so do NOT
subject it to strong vibrations or shock. Additionally, the heater unit reaches high temperatures
and is subjected to high voltages. So, maintenance services should be performed after the power
is off and the heater unit temperature has returned to normal room temperature.
For details, refer to IM11M12A01-21E “Heater Assembly“.
NOTE
If the heater strut assembly can not be removed because a screw has fused to its thread, one of
our service representatives can fix it.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
11-5
<11. Inspection and Maintenance>
16
A
14
14
11
10
14
12
35
A
13
15
24
8
9
24
7
5
4
6
3
24
2
1
23
TC +
View A-A
17
25
TC -
CELL +
34
28
26
18
29
19
1
CELL
2 3
TC
4 5
6
CJ
33
13
21
14
36
37
35
Figure 11.4
22
30
32
7
H T R
8
F11.4.ai
20
Exploded View of Detector (When pressure compensation specified)
Note: The parts marked by is not equipped with the types except the pressure compensation type.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<11. Inspection and Maintenance>
11-6
Replacement of heater strut assembly (ZR22G : Style S2 and after)
Refer to Figure 11.4 as an aid in the following discussion.
Remove the cell assembly (6), following Section 11.1.2, earlier in this manual. Open the terminal
box (16) and remove the three terminal connections – CELL +, TC + and TC -.Before disconnect
the HTR terminals, remove the terminal block screw (28). Keeping the other terminal remaining
to be connected. Disconnect the two HTR connections. (These terminals have no polarity.)
Remove the two screws (15) that fasten the cover (12) and slide it to the flange side.
Remove the four bolts (10) and terminal box (16) with care so that the already disconnected wire
will not get caught in the terminal box.
In case of the pressure compensation type detector, remove the screw (36) and the plate (37) on
the adapter (35). Remove the adapter (35), drawing out the wires of the heater strut assembly
(23) from it.
Loosen screw (19) until heater strut assembly (23) plate can be removed.
There’s no need to remove O-ring (18) which prevents screw (19) from coming out.
Pull out connector (13).
Loosen and remove the screw for the heater assembly fixation (8) with a special wrench (part no.
K9470BX or equivalent) and then remove the heater strut assembly (23) from the detector (24).
To reassemble the heater strut assembly, reverse the above procedure:
Insert the heater strut assembly (23) into the detector (24), while inserting the calibration pipe in
the detector (24) into the heater section in the heater strut assembly (23) as well as in the bracket
hole. Coat the screw for the heater assembly fixation (8) with grease (NEVER SEEZ: G7067ZA)
and tighten the screw for the heater assembly fixation (8) with a special tool (part no. K9470BX or
equivalent) with a tightening torque of 12N • m ± 10%.
Next, to install the O-rings (22) on the calibration gas and reference gas pipes, disassemble the
connector (13) in the following procedure:
First, remove the screw (25) and then remove the plate (17) and two caps (20). If the O-rings (22)
remains in the hole, pull them out from the back. Pass the heater and thermocouple lead wire
through the connector (13). Also, pass the calibration gas and reference gas pipes through the
opening of the connector (13). If the O-ring (22) fails, replace it with a new one.
Push the two caps (20) into the associated opening of the connector (13). Insert the plate (17),
aligning it with the groove of the cap (20), and tighten it with the screw (25). If you attempt to
insert the calibration gas and reference gas pipes into the connector (13) without disassembling
the connector (13), the O-ring may be damaged. Tighten the screw (19) in heater strut assembly
(23) until connector (13) can’t move.
Reassemble in reverse order to the above disassembly procedure.
The two wires with ceramic insulators from the heater strut assembly are heater wires, and the
single-core shielded wire is the cell signal + terminal; for the two-core shielded cable, the semitranslucent rubber-sheathed wire is the thermocouple + terminal, and the other wire is the
− terminal. (If the wires are marked, match the markings with those on the terminal board).
When installing the cell assembly (6), replace the metal O-ring (7) with a new one.
11.1.4
Replacement of Dust Filter
Set the dust filter (1) in place using a special pin spanner (with a pin 4.5 mm in diameter: part no.
K9471UX or equivalent). If a dust filter that has already been replaced once is used again, apply
grease (NEVER SEEZ: G7067ZA) to the threads of the dust filter.
11.1.5
Replacement of O-ring
The detector uses three different types of O-rings (14), (21), and (22). One O-ring alone (14), or
two O-rings (21) and (22) are used. (For a pressure compensating model, two O-rings are used
for individual uses. Two O-rings (21) and (22) are used for reference gas and calibration gas
sealing and require periodic replacement.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<11. Inspection and Maintenance>
Part No.
Description
(7)
K9470BJ
Metal O-ring
(14)
K9470ZS
O-ring with grease
(21)
(22)
K9470ZP
Two pairs of O-rings with grease
11.1.6
11-7
Cleaning the High Temperature Probe Adapter
CAUTION
Do not subject the probe of the High Temperature Probe Adapter (ZO21P-H-A) to shock. This
probe uses silicon carbide (SiC) which may become damaged if it is subjected to a strong shock
or thermal shock.
The high temperature detector is structured so that the gas to be measured is directed toward
the detector with the high temperature probe adapter. Therefore, if the probe or the sample gas
outlet clogs, a precise measurement is no longer possible because of no gas flow. If you use the
high temperature detector, you have to inspect it periodically and, if any part of it is significantly
clogged with dust, clean it.
Dust found sticking to the probe should be blown off. If any dust still remains after the blowing,
clean it with a metal rod, etc., inserted. In addition, if dust is found on the auxiliary ejector or
needle valve (choke) at the sample gas outlet, remove these parts from the high temperature
probe adapter and then clean them. To remove dust, blow air on them or rinse them with water.
11.1.7
Stopping and Re-starting Operation
When operation is stopped, take care of the followings so that the sensor of the detector cannot
become unused.
CAUTION
When operating an instrument such as boiler or industrial furnace is stopped with the zirconia
oxygen analyzer operation, moisture can condensate on the sensor portion and dusts may stick
to it.
If operation is restarted in this condition, the sensor which is heated up to 750°C firmly fixes the
dusts on itself. Consequently, the dusts can make the sensor performance much lower. If a large
amount of water is condensed, the sensor can be broken and will never be used.
To prevent the above nonconformity, take the following action when stopping operation.
(1) If possible, keep on supplying the power to converter and flowing reference gas to the sensor.
If this is impossible to do the above, remove the detector.
(2) If unavoidably impossible to supply the power and removing the detector, keep on the
following air at 600 ml/min into the calibration gas pipe.
When restarting operation, be sure to flow air, for 5-10 minutes, at 600 ml/min into the calibration
gas pipe before supplying power to the converter.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<11. Inspection and Maintenance>
11.2
11-8
Inspection and Maintenance of the
Converter
The converter does not require routine inspection and maintenance. If the converter does not
work properly, in most cases it probably comes from problems or other causes.
A dirty touchpanel should be wiped off with a soft dry cloth.
11.2.1
Replacing Fuses
The converter incorporates a fuse, as indicated in Figure 11.5. If the fuse blows out, replace it in
the following procedure.
CAUTION
• If a replaced fuse blows out immediately, there may be a problem in the circuit.
Go over the circuit completely to find out why the fuse has blown.
• This fuse is for protecting the main power supply circuit and does not provide overcurrent
protection for the heater temperature control circuit. For overcurrent protection circuitry, refer
to Section 12.1.2.2, Heater Temperature Failure.
Fuse
F11.5E.ai
Figure 11.5
Location of Fuse in the Converter
To replace the fuse, follow these steps:
(1) Turn off the power to the converter for safe replacement.
(2) Remove the fuse from its holder. With the appropriate flat-blade screwdriver that just fits the
holder cap slot (Figure 11.6), turn the fuse holder cap 90° counterclockwise.
By doing so, you can remove the fuse together with the cap.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<11. Inspection and Maintenance>
11-9
Cap
Flat-blade
screwdriver
Socket
Fuse
F11.6E.ai
Figure 11.6
Removing the Fuse
(3) Check the rating of the fuse and that it satisfies the following:
Maximum rated voltage: 250 V
Maximum rated current: 3.15 A
Type: Time-lag fuse
Standards: UL-, CSA- or VDE-approved
Part number: A1113EF
Place a new, properly rated fuse in the holder together with the cap, and push and turn the
cap clockwise 90° with the screwdriver to complete installation of the fuse.
11.2.2 Cleaning
Use a soft dry cloth to clean any part of the converter during inspection and maintenance.
11.2.3
Adjust LCD screen contrast
An LCD is built in the ZR402G converter. The contrast of this LCD is affected by its ambient
temperature. For this reason, the LCD is shipped, after adjusting the contrast so as to become
the most suitable in a room temperature(20-30°C). However, when display on the LCD is hard
to see, adjust the LCD contrast by change the resistance of the variable resistor; its position is
shown in Fig. 11.7
Variable register
Figure 11.7
11.3
Replacing Flowmeter in ZR40H Automatic
Calibration Unit
(1) Remove piping and wiring, and remove the ZR40H from the 2B pipe or wall mounting.
(2) Remove four M6 bolts between brackets.
(3) Remove piping extension
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<11. Inspection and Maintenance>
11-10
(4) Remove bolts holding flowmeter, and replace it. A white back plate (to make the float easy
to see) is attached. The end of the pin holding down the back plate must be on the bracket
side.
(5) Replace piping, and fix M6 bolts between brackets. *1
*1 : When disassembling and reassembling, mark original positions, and tighten an extra 5-10° when reassembling.
After tightening, do a liquid leakage test.
Flowmeter with back plate
Bolt fixes bracket
COM ON S N
Z RO
Connecter
F11.8E.ai
Figure 11.8
Flowmeter replacement
C
D
Two screws fix flowmeter
Connect piping pairs A-A', B-B', C-C', D-D'
A'
A
C'
B'
D'
B
F11.9E.ai
Figure 11.9
Fixing Flowmeter
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
12-1
<12. Troubleshooting>
12. Troubleshooting
This chapter describes errors and alarms detected by the self-diagnostic function of the
converter. It also explains inspections and remedies when other problems occur.
12.1
Displays and Remedies When Errors Occur
12.1.1
Error Types
An error occurs when an abnormality is detected in the detector or the converter, e.g., in the cell
(sensor), detector heater, or internal circuits of the the converter..
If an error occurs, the converter performs the following:
(1) Stops the supply of power to the heater in the detector to insure system safety.
(2) Causes an error indication in the display to start blinking to notify of an error generation
(Figure 12.1).
(3) Sends an output contact if the error is set up for Output contact setup for that contact
(refer to Section 8.4, Output Contact Setup).
(4) Changes the analog output status to the one set in Output hold setting.
(refer to Section 8.2, Output Hold Setting).
When the display shown in Figure 12.1 appears, pressing the error indication brings up a
description of the error (Figure 12.2). The content of errors that are displayed include those
shown in Table 12.1.
Tag:
Tag:
21.0
Error 2 :Heater temp.
Alarm 1 1 : T.C. voltage
%O2
17.42mA -Output1
17.42mA -Output2
Hold
Hold
F12.1E.ai
F12.2E.ai
Figure 12.1 Figure 12.2
Table 12.1
Error Code
Types of Errors and Reasons for Occurrence
Error Type
Occurrence Conditions
Error1
Cell voltage failure
The cell (sensor) voltage signal input to the converter falls below
-50 mV.
Error2
Heater temperature failure
The heater temperature does not rise during failure warm-up, or it
falls below 730ºC or exceeds 780ºC after warm-up is completed.
Or this occurs if the TC+, TC- thermocouple terminals are wired to
converter with reverse (wrong) polarity.
Error3
A/D converter failure
The A/D converter fails in the internal electrical circuit in the
converter.
Error4
Memory failure
Data properly are not written into memory in the internal electrical
circuit in the converter.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
12.1.2
12-2
<12. Troubleshooting>
Remedies When an Error Occurs
Error1: Cell Voltage Failure
Error1 occurs when the cell (sensor) voltage input to the converter falls below -50 mV
(corresponding to about 200 vol%O2). The following are considered to be the causes for the cell
voltage falling below -50 mV:
(1) Poor contact in terminal connections between the converter and detector
(2) Breakage in wiring cable between the converter and the detector
(3) Damage or deterioration of the sensor assembly
(4) Continuity failure between the sensor assembly electrode and the contact
(5) Wiring failure inside the detector
(6) Abnormality in electrical circuits in the converter
Turn off power to the converter.
Is there any breakage
Yes
or poor contact in the wiring to the
converter terminals?
Replace the damaged
portions.
No
Is there any breakage
Yes
or poor contact in the wiring to the
detector terminals?
Replace the damaged
portions.
No
Is the sensor extremely
dirty, corroded or broken?
Yes
See Section 11.1.2 for the
replacement procedure.
No
Is there any
disconnection or poor continuity
in the wiring between the
detector and
converter?
Replace the sensor
assembly.
Yes
Replace the wiring cable.
No
Is an error indicated
on the display?
No
Yes
Replace the sensor
assembly and temporarily
place the analyzer in the
operating status.
Examine the wiring connection to converter terminals
"CELL+" and "CELL-" Also, check the wiring
connection in the repeater terminal box if it is used.
Examine the wiring connection status to
detector terminals 1 and 2. Also, check that
terminals and cable conductors are not corroded.
Remove the sensor assembly from the detector
and check for the presence of corrosion that may
cause a poor contact between the electrode and
the contact; also, check for dirt. A sensor assembly
in which no abnormality is found may be used again.
However, be sure to use a new metal O-ring and
contact even in such a case.
Remove the wiring conductors from detector
terminals 1 and 2 and short out the removed
conductors. Measure the resistance of these
wiring conductors on the converter side.
The resistance value is normal if it indicates
10 Ω or less.
Turn on the power to the converter and temporarily
place the analyzer in the operating status.
The analyzer operates normally.
See Section 11.1.2 for the replacement
procedure.
End. Carry out calibration.
No
Yes
A failure in the detector or
the converter is suspected.
Contact Yokogawa.
F12.1.2E.ai
Error2: Heater Temperature Failure
This error occurs if the detector heater temperature does not rise during warm-up, falls below 730
°C after warm-up, or exceeds 780°C. When Error2 occurs, Alarm 10 (cold junction temperature
alarm) or Alarm 11 (thermocouple voltage alarm) may be generated at the same time. Be sure to
press the error indication to get the error description and confirm whether or not these alarms are
being generated simultaneously.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<12. Troubleshooting>
12-3
If Alarm 10 is generated simultaneously, a failure in the cold junction system on the detector
terminal block is suspected. In this case, follow the procedure according to troubleshooting for
Alarm 10 in 12.2.2.5.
If Alarm 11 is generated simultaneously, a failure in the thermocouple system located in the
detector heater is suspected. In this case, follow the procedure according to troubleshooting for
Alarm 11 in 12.2.2.6.
If this failure occurs immediately after the power is supplied, the polarity at thermocouple input
connection (TC+, TC-) on the converter may be reversed. Check the connection from the
detector.
Causes considered for cases where Error2 occurs independently are shown below.
(1) Faulty heater in the detector (heater wire breakage)
(2) Faulty thermocouple in the detector
(3) Faulty cold junction sensor located at the detector terminal block.
(4) Failure in electrical circuits inside the converter
(5) Heater temperature control overcurrent limiting triggered.
(6) TC+ ,TC- thermocouple terminals wired to detector with reverse (wrong) polarity.
Overcurrent protection is triggered if there are problems in the heater wiring. When the protective
circuit is triggered, the internal fuse blows and the heater is disconnected, resulting in Error 2
(temperature failure).
(1) Turn off power to the converter.
(2) Remove the cable from terminals 7 and 8 of the detector and measure the resistance value
between these terminals. The heater unit is normal if the resistance is lower than about
90 Ω. If the resistance value is higher, failure of the heater unit is suspected. In this case,
replace the heater unit (refer to Section 11.1.3, Replacement of the Heater Unit). In addition,
check that the wiring resistance between the converter and detector is 10 Ω or less.
(3) Ensure that TC+ terminal (terminal 3 in detector) is connected to converter TC+ terminal,
and TC- terminal (terminal 4) is connected to converter TC- terminal.
(4) Remove the wiring from terminals 3 and 4 of the detector and measure the resistance
value between these terminals. The thermocouple is considered normal if the resistance
value is 5 Ω or less. If the value is higher than 5 Ω, it may indicate that the thermocouple
wire has broken or is about to break. In this case, replace the heater unit (refer to Section
11.1.3, Replacement of the Heater Unit). Also, check that the wiring resistance between the
converter and detector is 10 Ω or less.
(5) Even if items (2) to (4) are normal, the heater overcurrent protection fuse may have blown.
Check for wiring problems such as the following:
(a) Heater terminals shorted.
(b) Heater terminal(s) shorted to ground.
(c) Heater terminals shorted to power supply.
If the internal fuse blows, this cannot be replaced by the user. Contact your Yokogawa service
representative.
NOTE
Measure the thermocouple resistance value after the temperature difference between the
detector tip and the ambient atmosphere has decreased to 50°C or less. If the thermocouple
voltage is large, accurate measurement cannot be achieved.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
12-4
<12. Troubleshooting>
Error3: A/D Converter Failure/Error4: Writing-to-memory Failure
• A/D Converter Failure
It is suspected that a failure has occurred in the A/D converter mounted in the electrical circuits
inside the converter.
• Writing-to-memory Failure
It is suspected that a failure has occurred in an operation writing to the memory (EEPROM)
mounted in the electrical circuits inside the converter.
Turn off the power to the converter once and then restart the converter. If the converter
operates normally after restarting, an error might have occurred due to a temporary drop in the
voltage (falling below 85 V, the least amount of voltage required to operate the converter) or a
malfunction of the electrical circuits affected by noise. Check whether or not there is a failure in
the power supply system or whether the converter and detector are securely grounded.
If the error occurs again after restarting, a failure in the electrical circuits is suspected.
Consult the service personnel at Yokogawa.
12.2
Displays and Remedies When Alarms are
Generated
12.2.1
Alarm Types
When an alarm is generated, the alarm indication blinks in the display to notify of the alarm
(Figure 12.3). Pressing the alarm indication displays a description of the alarm.
Alarms include those shown in Table 12.2.
Tag:
Tag:
21.0
Alarm 6 :Zero conc.ratio
%O2
17.42mA -Output1
17.42mA -Output2
F12.3E.ai
F12.4E.ai
Figure 12.3 Figure 12.4
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<12. Troubleshooting>
Table 12.2
Alarm
Code
12-5
Types of Alarms and Reasons for Occurrence
Alarm Type
Occurrence Conditions
Alarm 1
Oxygen concentration Measured oxygen concentration value exceeds or falls below the preset
alarm
alarm limits. (refer to Section 8.3, Setting Alarms).
Alarm 2
Humidity alarm
Occurs when the humidity to be measured exceeds or falls below the set
alarm points (refer to Section 8.3, “Alarm Setting”).
Alarm 3
Mixing-ratio alarm
Occurs when the mixing ratio to be measured exceeds or falls below the
set alarm points (refer to Section 8.3, “Alarm Setting”).
Alarm 6
Zero-point calibration
coefficient alarm
In automatic or semi-automatic calibration, zero correction ratio is outside
the range of 100 ±30%. (refer to Section 9.1.4, Compensation).
Alarm 7
Span-point calibration
coefficient alarm
Span correction ratio is outside the range of 0±18% (refer to Section 9.1.4,
Compensation).
Alarm 8
EMF stabilization
time-up alarm
In automatic or semi-automatic calibration, cell voltage does not stabilize
after calibration time is up.
Alarm 9
Exhaust gas
temperature alarm
When “mA-input” is selected in the Exhaust gas setup display, this alarm
occurs if the exhaust gas temperature exceeds the set alarm values (refer
to Section 8.6.4, “Setting Measurement Gas Temperature and Pressure”).
Alarm 10
Cold junction
temperature alarm
Temperature of the cold junction placed in the detector terminal box falls
below -25ºC or exceeds 155ºC.
Alarm 11
Thermocouple
voltage alarm
Generated when thermocouple voltage exceeds 42.1 mV (about 1020 ºC)
or falls below -5 mV (about -170ºC).
Alarm 12
Input current alarm
When “mA-input” is selected in the Exhaust gas setup display, this alarm
occurs if the input current from the temperature transmitter is outside from
3.2 to 21.6 mA.
Alarm 13
Battery low alarm
Internal battery needs replacement.
If an alarm is generated, actions such as turning off the heater power are not carried out. The
alarm is cancelled when the cause of the alarm is removed. However, Alarm 10 and/or Alarm 11
may be generated concurrently with Error2 (heater temperature error). In this case, the operation
when the error occurs has priority.
If the power to the converter is turned off after an alarm is generated and the converter is restarted
before the cause of the alarm has been removed, the alarm will be generated again. However,
Alarms 6, 7, and 8 (alarms related to calibration) are not generated unless calibration is executed.
12.2.2
Remedies When Alarms are Generated
Alarm 1, Alarm 2, and Alarm 3: Oxygen Concentration Alarm, Humidity
Alarm and Mixing Ratio Alarm
This alarm is generated when a measured value exceeds an alarm set point or falls below it. For
details on these alarms, see Section 8.3, Setting Alarms, in the chapter on operation.
Alarm 6: Zero-point Calibration Coefficient Alarm
In automatic or semi-automatic calibration, this alarm is generated when the zero correction ratio
is out of the range of 100 ± 30% (refer to Section 9.1.4, Compensation).
The following are possible causes of this alarm.
(1) The zero-gas oxygen concentration does not agree with the value of the zero-gas
concentration set in Calibration setup. Otherwise, the span gas is used as the zero gas.
(2) The flow rate of zero gas is out of the specified flow (600 ± 60 ml/min).
(3) The sensor assembly is damaged and so cell voltage is not normal.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<12. Troubleshooting>
12-6
(1) Check that the following have been set up correctly. If not, correct them. Then, recalibrate.
a. Check the preset zero gas concentration on the Calibration setup display. The displayed
concentration value has agreed with the concentration of the zero gas actually used.
b. The piping for calibration gases has been constructed so that the zero gas does not leak.
(2) If the alarm is not generated during the recalibration, improper calibration conditions are
considered as the cause of the alarm generated in the previous calibration. In this case, no
particular restoration is necessary.
(3) If the alarm is generated again during the recalibration, deterioration of or damage to the
sensor assembly is considered as the cause of the alarm. It is necessary to replace the cell
with a new one. Before replacement, carry out the following.
Check the cell voltages when the zero gas and span gas are introduced.
a. From the Basic panel display, move to the Detailed-data display of the channel where the
alarm is being generated.
b. When the [▼] key is pressed once, the cell voltage should be indicated on the top line
(Figure 12.5).
c. Check whether or not the displayed cell voltage is very different from the theoretical value
at the oxygen concentration of the relevant gas. See Table 12.3 for the theoretical cell
voltages. Although the tolerance to the theoretical value cannot be generally specified, a
reasonable one may be approximately ±10 mV.
Table 12.3
Oxygen Concentration and Cell Voltage
Oxygen concentration
(vol%O2)
Cell voltage
(mV)
1%
67.1
21%
0
(4) Check whether the deterioration of or damage to the sensor assembly that caused the
alarm has occurred suddenly during the current calibration in the following procedure:
a. Call up the Detailed-data Display.
b. Use the [Page Scroll] key to check Calibration Data (Figure 12.6). The span and zero
correction ratios of the last ten calibration can be checked here. By checking these data,
whether the sensor deterioration has occurred suddenly or gradually can be determined.
Tag:
Tag:
Cell voltage:
0 . 6 mV
4 2 . 1 mV
Thermo voltage:
C.J.resistance: 1 1 8 1 . 4 Ω
-4Ω
Cell resistance:
0.24
Soft.rev.:
Calibration data:
1: 2 0 0 0 / 07 / 07
Span gas ratio:
Zero gas ratio:
2: 2 0 0 0 / 07 / 07
Span gas ratio:
Zero gas ratio:
Hold
F12.5E.ai
Figure 12.5
Detailed Data Display
1 4 : 57
-2.9%
70.3%
14:54
-2.9%
89.7%
Hold
Figure 12.6
Calibration History
(5) If the sensor assembly has deteriorated suddenly, the check valve that prevents moisture in
the furnace from entering into the calibration pipes may have malfunctioned. If the furnace
gas flows into calibration lines, the gas is cooled and thus condensation develops and
accumulates in the pipe. During calibration the condensate is carried with the calibration gas
and blow onto the sensor assembly, whereby the cell is cooled quickly. This results in the
failed sensor assembly.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<12. Troubleshooting>
12-7
(6) If the sensor assembly has deteriorated gradually, check the condition of the sensor
assembly following the procedure below.
a. Use the [Page Scroll] key to check Cell Resistance. It should be 200 V or less if the cell
(sensor) is new. On the other hand, if the cell (sensor) is approaching the end of its service
life, it will be 3 to 10 kV.
b. Use the [Page Scroll] key to check Cell Robustness. It should say “Life > 1 year” if the cell
(sensor) is in good condition.
Tag:
Span gas ratio:
-2.9%
Zero gas ratio:
70.3%
Response time:
0s
Cell robustness: life< 1 month
Cell temperature:
1 0 6 7 °C
C.J.temperature:
4 7 °C
Hold
F12.7E.ai
Figure 12.7
Alarm 7: Span-point Calibration Coefficient Alarm
In automatic or semi-automatic calibration, this alarm is generated when the span correction ratio
is out of the range of 0 ± 18% (refer to Section 9.1.4, Compensation).
The following are suspected as the cause:
(1) The oxygen concentration of the span gas does not agree with the value of the span gas set
in Calibration setup.
(2) The flow of the span gas is out of the specified flow value (600 ± 60 ml/min).
(3) The sensor assembly is damaged and the cell voltage is abnormal.
(1) Confirm the following and carry out calibration again: If the items are not within their proper
states, correct them.
a. If the display “Span gas conc” is selected in Calibration setup, the set point should agree
with the concentration of span gas actually used.
b. The calibration gas tubing should be constructed so that the span gas does not leak.
(2) If no alarm is generated as a result of carrying out re-calibration, it is suspected that
improper calibration conditions were the cause of the alarm in the preceding calibration.
In this case, no specific restoration is necessary.
(3) If an alarm is generated again as a result of carrying out re-calibration, deterioration of or
damage to the cell (sensor) is suspected as the cause of the alarm. Replacement of the
cell with a new one is necessary. However, before replacement, carry out the procedure
described in step (3) and later of in
Section 12.2.2.2, Alarm 6: Zero-point Calibration Coefficient Alarm.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<12. Troubleshooting>
12-8
Alarm 8: EMF Stabilization Time-up Alarm
This alarm is generated if the sensor (cell) voltage has not stabilized even after the calibration
time is up for the reason that the calibration gas (zero gas or span gas) has not filled the sensor
assembly of the detector.
(1) The flow of the calibration gas is less than normal (a specified flow of 600 ± 60 ml/min).
(2) The length or thickness of the calibration gas tubing has been changed (lengthened or
thickened).
(3) The measuring gas flows toward the tip of the probe.
(4) The sensor (cell) response has deteriorated.
(1) Carry out calibration by passing the calibration gas at the specified flow (600 ± 60 ml/min)
after checking that there is no leakage in the tubing.
(2) If calibration is carried out normally, perform a steady operation without changing the
conditions. If the error occurs again, check whether or not the reason is applicable to the
following and then replace the sensor assembly.
• A lot of dust and the like may be sticking to the tip of the detector probe. If dust is found,clean
the probe (see Section 11.1.1).
In addition, if an error occurs in calibration even after the sensor assembly is replaced, the
influence of sample gas flow may be suspected. Do not let the sample gas flow toward the
tip of the detector probe, for example, by changing the mounting position of the detector.
Alarm 9: Exhaust Gas Temperature Alarm
When “mA-input” is selected in the Exhaust gas setup display, this alarm occurs if the exhaust
gas temperature exceeds the set alarm values.
The following are probable causes:
(1) The temperature transmitter output range does not meet the analyzer Exhaust gas setup (if
this alarm occurs when the equipment starts up).
(2) Thermocouple(s) connected to the temperature transmitter may be defective
(disconnected).
(3) Temperature transmitter may be defective.
(4) Exhaust gas temperature may exceed the set alarm value.
(1) Check that the temperature transmitter output temperatures at 4 and 20 mA meet the
temperatures at 4 mA and 20 mA set with this equipment.
(2) Check that an actual exhaust gas is normal.
Press the detailed data display key to display and check that the exhaust gas temperature
in the displayed detailed data is normal. If this value is outside the temperature transmitter’s
normal output, the thermocouple(s) connected to the temperature transmitter may be damaged
(disconnected). See the applicable temperature transmitter instruction manual for solving
problems.
Alarm 10: Cold Junction Temperature Alarm
This alarm is generated when the temperature of the cold junction located at the terminal block of
the detector falls below -25°C or exceeds 155°C. Check the following:
Display “C.J.temperature” in the Detailed-data display. If “C.J.temperature” is indicated as 200°C
or -50°C, the following can be considered.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<12. Troubleshooting>
12-9
(1) Breakage of the cold junction signal wires between the converter and the detector, or the
cable is not securely connected to the connecting terminals.
(2) The positive and negative poles of the cold junction signal wiring are shorted out in the
wiring extension or at the connection terminals.
(3) A failure of the cold junction temperature sensor located at the detector terminal block
occurred.
(4) A failure of the electrical circuits inside the converter occurred.
If “C.J.temperature” exceeds 150°C or falls below -20°C, the following can be considered.
• The temperature of the detector terminal block is out of the operating temperature range
(-20°C to 150°C).
• A failure of the cold junction temperature sensor located at the detector terminal block
occurred.
• A failure of the electrical circuits inside the converter occurred.
Before proceeding to the following troubleshooting procedure, examine whether or not the
temperature of the detector terminal block is out of the operating temperature range.
The operating temperature range varies with the type of detector. If the detector terminal block
is out of its operating temperature range, take the measure to lower the temperature, such as
situating it so that it is not subjected to radiant heat.
The case where the Model ZR22 Detector is used:
(1) Stop the power to the converter.
(2) Remove the wiring from terminals 5 and 6 of the detector and measure the resistance
between these terminals. If the resistance value is out of the range of 1 to 1.6 kΩ, the cold
junction temperature sensor is considered to be faulty. Replace that temperature sensor
with a new one.
(3) If the resistance value is within the above range, the cold junction temperature sensor
seems to be normal. Check whether or not the cable is broken or shorted out, and whether
the cable is securely connected to the terminals. Also, check that the resistance of the wiring
between the converter and detector is 10 Ω or less.
(4) If there is no failure in the wiring, the electrical circuits inside the converter may possibly fail.
Contact the service personnel at Yokogawa.
The case where the Model ZO21D Detector is used:
(1) Without stopping the power to the converter, remove the wiring from terminals 5 and 6 of
the detector and measure the voltage between these terminals. If the voltage between the
terminals is out of the range of 0.4 to 0.7 V, the cold junction temperature sensor seems to
be faulty. Replace the cold junction temperature sensor.
(2) If the voltage between the terminals is within the above range, the cold junction temperature
sensor seems to be normal. Check whether or not the cable is broken or shorted out, and
whether the cable is securely connected to the terminals. Also, check that the resistance of
the wiring between the converter and detector is 10 Ω or less.
(3) If there is no failure in the wiring, the electrical circuits inside the converter may possibly fail.
Contact the service personnel at Yokogawa.
CAUTION
The operating temperature range of the Model ZO21D Detector is -10°C to 80°C (except for the
high temperature detector ZO21D-H). Since a cold junction temperature alarm for this analyzer
is not generated until the temperature exceeds 155°C, if the Model ZO21D Detector is used, be
careful in controlling the ambient temperature of the terminal block.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<12. Troubleshooting>
12-10
Alarm 11: Thermocouple Voltage Alarm
This alarm is generated when the emf (voltage) of the thermocouple falls below -5 mV (about
-170°C) or exceeds 42.1 mV (about 1020°C). Whenever Alarm 11 is generated, Error2 (heater
temperature failure) occurs.
(1) Breakage of the heater thermocouple signal wire between the converter and the detector
occurred, or the cable is not securely connected to the connecting terminals.
(2) The positive and negative poles of the heater thermocouple signal wiring are shorted out in
the wiring extension or at the connection terminals.
(3) A failure of the thermocouple at the detector heater assembly occurred.
(4) A failure of the electrical circuits inside the converter occurred.
(1) Stop the power to the converter.
(2) Remove the wiring from terminals 3 and 4 of the detector and measure the resistance
between these terminals. If the resistance value is 5 Ω or less, the thermocouple seems to
be normal. If it is higher than 5 Ω, it may indicate the possibility that the thermocouple has
broken or is about to break. In this case, replace the heater unit (refer to Section 11.1.3,
Replacement of the Heater Unit).
CAUTION
Measure the thermocouple resistance value after the difference between the detector tip
temperature and ambient temperature falls to 50°C or less. If the thermocouple voltage is large,
accurate measurement cannot be achieved.
(3) If the thermocouple is normal, check whether or not the wiring cable is broken or shorted
out, and also whether the wiring cable is securely connected to the terminals.
Also check that the wiring resistance between the converter and the detector is 10 Ω or less.
(4) If there is no failure in the wiring, the electrical circuits inside the converter may possibly fail.
Contact the service personnel at Yokogawa.
Alarm 12: Input Current Alarm
When “mA-input” is selected in the Exhaust gas setup display, this alarm occurs if the current
input from the temperature transmitter goes outside he range from 3.2 to 21.6 mA.
If this alarm occurs simultaneously with Alarm 9: Exhaust Gas Temperature Alarm, first solve
the problem for Alarm 9. If Alarm 12 occurs independently, the cable connection between this
equipment and the temperature transmitter may be improper (disconnected).
(1) Check that cable connections (including connecting lugs) are proper.
(2) If the cable connections are correct, display the exhaust gas temperature and check that it
matches the temperature transmitter’s temperature signals. If mismatched, check whether
the transmitter output range meets the Exhaust gas setup of this equipment.
(3) If the range setting is correct, the analyzer electronics may be defective. In such a case,
contact your local Yokogawa service or sales representative.
Alarm 13: Battery Low Alarm
An internal battery is used as backup for the clock. After this alarm occurs, removing power from
the instrument may cause the clock to stop but should not affect stored parameters. The internal
clock is used for blow back scheduling; if you use this then after a battery alarm occurs (until the
battery is replaced) be sure to check / correct the date and time every time you turn on the power.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<12. Troubleshooting>
12-11
When the battery low alarm occurs, remember that the battery cannot be replaced by the user.
Contact your Yokogawa service representative.
NOTE
Battery life varies with environmental conditions.
* If power is applied to the instrument continuously, then the battery should not run down, and
life is typically about ten years. However the battery will be used during the time interval
between shipment from the factory and installation.
* If power is not applied to the instrument, at normal room temperatures of 20 to 25°C then
battery life is typically 5 years, and outside this range but within the range -30 to +70°C then
battery life is typically 1 year.
12.3
Countermeasures When Measured Value
Shows Error
The causes that the measured value shows an abnormal value is not always due to instrument
failures. There are rather many cases where the causes are those that measuring gas itself
is in abnormal state or external causes exist, which disturb the instrument operation. In this
section, causes of and measures against the cases where measured values show the following
phenomena will be described.
(1) The measured value is higher than the true value.
(2) The measured value is lower than the true value.
(3) The measured value sometimes shows abnormal values.
12.3.1
Measured Value Higher (Lower for Humidity Analyzer)
Than True Value
(1) The measuring gas pressure becomes higher.
The measured oxygen concentration value X (vol%O2) is expressed as shown below,
when the measuring gas pressure is higher than that in calibration by ∆p (kPa).
X=Y [ 1+ (∆p/101.30) ]
where Y: Measured oxygen concentration value at the same pressure as in calibration
(vol%O2).
Where an increment of the measured value by pressure change cannot be neglected,
measures must be taken.
Investigate the following points to perform improvement available in each process.
• Is improvement in facility’s aspect available so that pressure change does not occur?
• Is performing calibration available under the average measuring gas pressure
(internal pressure of a furnace)?
(2) Moisture content in a reference gas changes (increases) greatly.
If air at the detector installation site is used for the reference gas, large change of moisture
in the air may cause an error in measured oxygen concentration value (vol%O2).
When this error is not ignored, use a gas in which moisture content is constant such as
instrument air in almost dry condition as a reference gas.
In addition, change of moisture content in exhaust gas after combustion is also considered
as a cause of error. However, normally this error is negligible.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<12. Troubleshooting>
12-12
(3) Calibration gas (span gas) is mixing into the detector due to leakage.
If the span gas is mixing into the detector due to leakage as a result of failure of the valve
provided in the calibration gas tubing system, the measured value shows a value a little
higher than normal.
Check valves (needle valves, check valves, solenoid valves for automatic calibration,
etc.) in the calibration gas tubing system for leakage. For manual valves, check them after
confirming that they are in fully closed states. In addition, check the tubing joints for leakage.
(4) The reference gas is mixing into the measuring gas and vice versa.
Since the difference between oxygen partial pressures on the sensor anode and cathode
sides becomes smaller, the measured value shows a higher value.
An error which does not appear as the Error1 may occur in the sensor. Sample gas and/or
the reference gas may be leaking. Visually inspect the sensor. If any crack is found, replace
the sensor assembly with a new one.
Referring to Section 11.1.3, check if there is a problem with the sensor mounting. If things
are abnormal but there is no error indication, then maybe the sensor assembly isn’t properly
secured, the O-ring seal is dirty, the measurement gas and ratio gas are leaking into each
other (if measurement gas pressure is high, then it may leak into ratio gas, and vice versa);
as oxygen partial pressure difference between ratio gas and measurement gas is small, the
oxygen concentration in measured gas will read high and the humidity value will read low.
Replace the sensor assembly as shown in 11.1.3, and be sure to replace the metal O-ring
with a new one.
If the sensor is cracked, the symptoms will be the same – low humidity measured value.
Perform a visual inspection, and – if a crack is found – replace the sensor assembly as
described in 11.1.3.
NOTE
Data such as cell robustness displayed in the Detailed-data display should also be used for
deciding sensor quality as references.
12.3.2
Measured Value Lower (Higher for Humidity Analyzer)
Than True Value
(1) The measuring gas pressure becomes lower.
Where an increment of the measured value due to pressure change cannot be neglected,
take measures referring to subsection 12.3.1 (1).
(2) Moisture content in a reference gas changes (decreases) greatly.
If air at the detector installation site is used for the reference gas, large change of moisture
content in the air may cause an error in measured oxygen concentration value (vol%O2) or
humidity value (vol% H2O or kg/kg).
When this error is not ignored, use a gas in which moisture content is constant such as
instrument air in almost dry condition as a reference gas.
In addition, change of moisture content in exhaust gas after combustion is also considered
as a cause of error. However, normally this error is negligible.
(3) Calibration gas (zero gas) is mixed into the detector due to leakage.
If the zero gas is mixed into the detector due to leakage as a result of failure of the valve
provided in the calibration gas tubing system, the measured value shows a value a little
lower than normal.
Check valves (needle valves, check valves, solenoid valves for automatic calibration,
etc.) in the calibration gas tubing system for leakage. For manual valves, check them after
confirming that they are in a fully closed situation.
(4) Combustible components exist in the sample gas. If combustible components exist in the
sample gas, they burn in the sensor and thus oxygen concentration decreases.
Check that there are no combustible components.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
<12. Troubleshooting>
12-13
(5) Temperature of the detector cell reaches 750°C or more.
12.3.3
Measurements Sometimes Show Abnormal Values
(1) Noise may be mixing in with the converter from the detector output wiring.
Check whether the converter and detector are securely grounded.
Check whether or not the signal wiring is laid along other power cords.
(2) The converter may be affected by noise from the power supply.
Check whether or not the converter power is supplied from the same outlet, switch, or
breaker as other power machines and equipment.
(3) Poor wiring contact
If there is poor contact in the wiring, the sensor voltage or thermocouple emf (voltage) may
vary due to vibration or other factors.
Check whether or not there are loose points in the wiring connections or loose crimping
(caulking) at the crimp-on terminal lugs.
(4) Combustible components in the sample gas may be getting into the sensor.
If the combustible components show signs of dust, the abnormality may be improved by
mounting a dust-removing filter K9471UA.
(5) There may be a crack in the sensor or leakage at the sensor-mounting portion.
If the indication of concentration varies in synchronization with the pressure change in the
furnace, check whether or not there is a crack in the sensor or whether the sensor flange is
sticking tightly to the probe-attaching face with the metal O-ring squeezed.
(6) There may be leakage in the calibration gas tubing
In the case of a negative furnace inner pressure, if the indication of concentration varies with
the pressure change in the furnace, check whether or not there is leakage in the calibration
gas tubing.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
Blank Page
Customer
Maintenance
Parts List
Model ZR22G
Zirconia Oxygen/Humidity Analyzer, Detector
(Separate type)
1
ZR22A
Heater Assembly
7
2
3
6
5
10
4
Item
1
2
3
4
5
6
7
8
9
10
11
9
8
11
Part No.
Qty
K9471UA
--G7109YC
K9470BK
E7042DW
1
4
--K9470ZF
K9470ZG
E7042BR
K9470BM
1
K9473AN
--ZR01A01-01
ZR01A01-02
ZR01A01-05
1
1
ZR01A01-10
E7042BS
K9470BJ
E7042AY
--K9470ZK
K9470ZL
4
1
1
1
1
1
1
ZR22G.ai
Description
Dust Filter (Option)
Bolt
(M5x12, SUS316 stainless steel)
(M5x12, inconel) for Option code "/C"
Washer (SUS316 stainless steel)
Bolts and Washers
G7109YC x 4 + E7042DW x 4
K9470BK x 4 + E7042DW x 4 for Option code "/C"
Plate
Pipe
Pipe for Option code "/C"
Cell Assembly
1 piece (E7042UD)
2 pieces
5 pieces
10 pieces
Contact
Metal O-ring
Filter
Calibration Tube Assembly
Cal. Gas Tube Assembly
Cal. Gas Tube Assembly for Option code "/C"
All Rights Reserved, Copyright © 2000, Yokogawa Electric Corporation.
Subject to change without notice.
CMPL 11M12A01-02E
1st Edition : Aug. 2000 (YK)
9th Edition : Feb. 2016 (YK)
Customer
Maintenance
Parts List
Model ZR402G
Zirconia Oxygen Analyzer/High Temperature
Humidity Analyzer,Converter
1
Hood for ZR402G
ZR402G
2
Item
Part No.
Qty
Description
1
A1113EF
1
Fuse (3.15A)
2
K9471UF
1
Hood
All Rights Reserved, Copyright © 2000, Yokogawa Electric Corporation.
Subject to change without notice.
CMPL 11M12C01-01E
1st Edition : Jan. 2000 (YK)
3rd Edition : Feb. 2011 (YK)
Customer
Maintenance
Parts List
Model ZO21P-H
Zirconia Oxygen Analyzer
High Temperature Probe Adaptor
4
5,6
7
3
8,9,10
11
ZR22G
2
1
Item
Part No.
Qty
Description
Item
Part No.
Qty
Description
1
K9292TP
E7046CF
K9292TQ
E7046CG
E7046CH
1
1
1
1
1
Probe (SiC, L=0.5m)
Probe (SiC, L=0.6m)
Probe (SiC, L=0.7m)
Probe (SiC, L=0.8m)
Probe (SiC, L=0.9m)
2
E7046FA
E7046FE
E7046FK
E7046FD
E7046FC
1
1
1
1
1
Probe Adaptor (for JIS 5K-50-FF)
Probe Adaptor (for ANSI CLASS150-4-RF)
Probe Adaptor (for DIN PN10-DN50-A)
Probe Adaptor (for JIS 10K-100-FF)
Probe Adaptor (for JIS 10K-80-FF)
E7046AL
E7046BB
K9292TV
E7046CR
K9292TW
1
1
1
1
1
Probe (SiC, L=1.0m)
Probe (SiC, L=1.5m)
Probe (SUS, L=0.5m)
Probe (SUS, L=0.6m)
Probe (SUS, L=0.7m)
E7046FB
E7046FG
E7046FF
E7046FJ
E7046FH
1
1
1
1
1
Probe Adaptor (for JIS 10K-65-FF)
Probe Adaptor (for ANSI CLASS150-3-RF)
Probe Adaptor (for ANSI CLASS150-2 1/2-RF)
Probe Adaptor (for JPI CLASS150-4-RF)
Probe Adaptor (for JPI CLASS150-3-RF)
E7046CS
E7046CT
E7046AP
E7046AQ
1
1
1
1
Probe (SUS, L=0.8m
Probe (SUS, L=0.9m
Probe (SUS, L=1.0m
Probe (SUS, L=1.5m
3
4
5
6
7
E7046FQ
E7046FN
Y9825NU
Y9800WU
G7073XL
1
1
4
8
1
Gasket
Plate
Bolt
Washer
Gasket
8
9
10
11
Y9630RU
Y9121BU
Y9120WU
Y9801BU
4
4
4
4
Bolt
Nut
Washer
Nut
All Rights Reserved, Copyright © 2000, Yokogawa Electric Corporation.
Subject to change without notice.
CMPL 11M03B01-10E
1st Edition : Dec. 2000 (YK)
8th Edition : Sep. 2012 (YK)
Customer
Maintenance
Parts List
E7046EC/E7046EN
Zirconia Oxygen Analyzer
Auxiliary Ejector Assembly (for Model ZO21P-H)
1
3
2
4
5
6
7
Item
Part No.
Qty
1
E7046EC
1
Auxiliary Ejector Assembly, Connection Rc1/4
E7046EN
1
Auxiliary Ejector Assembly, Connection 1/4 NPT
L9852CB
1
Needle Valve, Connection Rc1/4
G7016XH
1
Needle Valve, Connection 1/4NPT
E7046EK
1
Pressure Gauge, Connection Rc1/4
E7046EV
1
Pressure Gauge, Connection 1/4NPT
E7046ED
1
Ejector, Connection Rc1/4
E7046EP
1
Ejector, Connection 1/4NPT
E7046EF
1
Nozzle Assembly, Tube Connection Ø6хØ4 tube
E7046ER
1
Nozzle Assembly, Tube Connection 1/4 inch tube
6
G7031XA
1
Tee, Connection Rc1/4
7
E7046EJ
1
Reducing nipple, Connection R1/4
2
3
4
5
Description
All Rights Reserved, Copyright © 2000, Yokogawa Electric Corporation.
Subject to change without notice.
CMPL 11M03B01-05E
1st Edition : Dec. 2000 (YK)
8th Edition : Mar. 2013 (YK)
Customer
Maintenance
Parts List
Model ZR40H
Automatic Calibration Unit
for Separate type Zirconia Oxygen/Humidity
Analyzer (ZR22G + ZR402G)
1
LOCK
ZERO
LOCK
COMMON SPAN ZERO
AUTO CALIBRATION UNIT
MODEL
SUFFIX
ZR40H
STYLE
S1
SUPPLY 690kPa MAX.
AMB.TEMP-20 TO 558C
USED WITHZR402G
NO.
Made in Japan
REF.OUT
CAL.OUT
SPAN IN
ZERO IN
Item
Part No.
Qty
Description
6
K9473XC
1
Flowmeter
All Rights Reserved, Copyright © 2000, Yokogawa Electric Corporation.
Subject to change without notice.
CMPL 11M12A01-11E
1st Edition : Dec. 2000 (YK)
4th Edition : Mar. 2013(YK)
Customer
Maintenance
Parts List
Model ZO21S
Zirconia Oxygen Analyzer/ High Temperature
Humidity Analyzer, Standard Gas Unit
1
2
3
Item
Part No.
Qty
Description
1
———
1
Pump (see Table 1)
2
E7050BA
1
Zero Gas Cylinder (x6 pcs)
3
E7050BJ
1
Needle Valve
Table 1
Power
Pump
AC 100 V
110
115
E7050AU
AC 200 V
220
240
E7050AV
All Rights Reserved, Copyright © 2000, Yokogawa Electric Corporation.
Subject to change without notice.
CMPL 11M3D1-01E
1st Edition : Jan. 2000 (YK)
4th Edition : Mar. 2011 (YK)
i
Revision Information
Manual Title : Model ZR22G, ZR402G Separate type Zirconia Oxygen/Humidity Analyzer
Manual No. : IM 11M12A01-02E
Apr. 2017/12th Edition
Addition RoHS etc.(pages i, vi, 2-2)
Feb. 2016/11th Edition
Bound up with IM 11M12A01-03E. IM 11M12A01-03E is obsoleted version.
Whole review.
CMPL 11M12A01-02E revised to 9th edition.
Aug.2015/10th.Edition
Revised section
2.1
“General Specifications”: Added to “Standard Specifications”
2-24 “Stop Valve”: Changed of the weight and dimensions.
2-24 “Check Valve”: Changed of the weight.
Nov. 2014/9th Edtion
Revised section
2.1.1 "Standard Specifications": Added the C-tick, Safety and EMC conforming standards.
2.2.1 "ZR22G Genenal use Separate type Detector" Sample gas pressure: Correction of erroneous
description. (0.5 to 5 kpa → -0.5 to 5 kPa)
2.2.2 "ZO21R Probe Protector": Deleted the flange of the DIN standard and changed the dimension
of the insertion length.
2.3.2 "ZO21P High Temperature Probe Adapter" Model and Codes: Added of the option code.
2.3.2 "ZO21P High Temperature Probe Adapter" Table: Changed the some numerical value.
2.4.1 "Standard Specification": Deleted the C-tick, Safety and EMC conforming standards.
5.1
"General" Table 5.1: Deleted cable type.
5.1
"General" : Added "WARNING".
5.2,
"Wiring for Detector Output" Ambient temperature of the detector : Changed temperature.
5.2.1 "Cable Specifications": Chenged description.
5.2.2 "Connection to the Detector" (2):Changed range of the ambient temperature at the location of
the wire installation.
5.2.3 "Connection to the Converter"(2):Changed description.
5.3
"Wiring for Power to Detector Heater" Ambient temperature of the detector: Changed
temperature.
5.3.1 "Cable Specifications": Chenged description.
5.3.2 "Connection to Detector" (2):Changed range of the ambient temperature at the location of the
wire installation.
5.3.3 "Connection to Converter" (2): Changed description.
5.5.2, "Grounding Wiring" (2):Changed range of the ambient temperature at the location of the wire
installation.
CMPL 11M03B01-10E revised to 8th edition.
CMPL 11M03B01-05E revised to 8th edition.
CMPL 11M12A01-02E revised to 8th edition.
CMPL 11M12A01-11E revised to 4th edition.
Jan. 2012/8th Edition
Revised and Corrected over all
5.1
Table 5.1 “Cable specifications”: Deleted cable type.
Added “WARNING”
5.2
Added the heat resistant request of the cable.
Changed range of the ambient temperature at the location of the wire installation.
Sep. 2006/7th Edition
Revised section
2.7.6 “Air Set,” Part No. K9473XH or K9473XJ, Standard Specification:
Changed descriptions partly; “Air Set,”Part No. G7004XF or K9473XG,
Standard Specification:Changed descriptions partly;
2.7.8 “Cylinder Regulator Valve (Part No. G7013XF or G7014XF), Standard Specification”:
Changed descriptions partly and drawing.
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
ii
8.2.3 “Output Hold Setting”: Changed value in Figures 8.2 and 8.3.
8.2.4 “Default Values”: Changed value and description in Table 8.3.
8.6.1 “Setting the Date-and-Time”: Added Item in Figure 8.15.
8.6.4 “Setting Purging”: Added Item.
8.6.5 “Setting Passwords”: Changed section number.
10.3 “Operational Data Initialization”: Changed value in Table 10.5.
Revised section
2.3.2 “ZO21P-H Adapter for High Temperature Probe”: Changed the thickness of the gasket in the
drawing.
2.4.1 “Standard Specification,” Safety and EMC conforming standards: The following CAUTION
description should be added.
“Standard Specification,” Construction: “ NEMA 4 “ should read “NEMA 4X/IP66.”
2.4.2 “• Model and Codes,”: Added Note and Suffix Codes.
8.2
“Output Hold Setting,” Table 8.1.2 : Added Note.
8.2.3 “Output Hold Setting” To set the output hold, follow these setps: The following Note
description should be added.
Revised and Corrected over all
Apr. 2005/6th Edition
Revised section
Introduction Added description in DANGER, Added description regarding modification
1.2.1 “ System Components” Changed part numbers of air set in table
2.2.1 Changed terminal box paint colors
2.4.1 Changed safety and EMC conforming standards and paint colors
2.5.2 Changed Finish color
2.6
Added description Non CE Mark
2.7.6 “ Air Set” Changed part numbers and drawing of air set
4.2.1 “ Piping Parts for System 2” Change part numbers of air set in Table 4.2
4.3.1 “ Blow Back Piping” Changed part numbers of air set
4.4.1 “ Piping Parts for a System using Detector with Pressure Conpensation” Changed part
numbers of air set in Table 4.3
5.3.3 Added WARNING
11.2.1 Added instruction in Note
12.1.1 Added description in Error-2 of Table 12.1, Type of Errors and Reasons for Occurrence
12.1.2.1 Changed reference information
12.1.2.2 Added descriptions
12.2.1 Added Alarm 13 in Table 12.2, Types of Alarms and Reasons for Occurrence
12.2.2.7 Added Section 12.2.2.7, “Alarm 13: Battery Low Alarm”
Dec. 2003/5th Edition
Changes of related by ROM and Main Board Assembly changed.
July 2003/4th Edition
Style of model ZR22G and ZR22A changed to S2
Notation of flange specification unified
Dust guard protector, Airset added
CMPL 11M12A01-02E Cell some parts no. changed, revised to 5th edition.
CMPL 11M03B01-05E some parts no. changed, revised to 6th edition.
Sep. 2001/3rd Edition
Revised section
1.2
Model ZR22A Heater Assembly added
2.5.1 ZA8F Flow Setting Unit error corrected
2.7.9 Model ZR22A Heater Assembly added
8.5.1 Table 8.7 Input Contact Functions changed
10.1.5 “French” added to Language Selection
11.1.3 Reference document added to Replacement of the Heater Unit
Heater Assembly added to CMPL 11M12A01-02E
CMPL11M03B01-05E revised to 6th Edtion, some part No. changed.
CMPL11M03B01-10E revised to 5th Edtion, some part deleted.
Mar. 2001/2nd Edition
Revised section
1.1.3 “System 3” explanation changed
2.2
In “General use Separate type Detector”, some MS codes changed
2.4
Corrected errors in ZR402G External Dimensions figure, changed MS Code table
2.5.1 Changed reference gas pressure where check valves is used, style changed of ZA8F
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
iii
2.5.2 Added detail about ZR40H Automatic Calibration Unit
2.7.3 Added items to figure of Auxiliary Ejector for High Temperature
3.2.2 Corrected Figure 3.6 Panel Cutout, Figure 3.13 Mounting Holes
3.5
Added to 3.5 Installation of ZR40H Automatic Calibration Unit
6.1
Added Filter to 6.1 ZR22G Detector
6.3
Added Names and Functions to 6.3 ZR40H Automatic Calibration Unit
7.2
Changed reference gas pressure where check valve is used
7.12.2.1 Changed reference gas pressure where check valve is used
10.1.1.6 Corrected Table 10.1
10.6.1 Changed reference gas pressure where check valve is used
11.1 Added Filter to 11.1 Inspection and Maintenance of the Detector Added Filter to
CMPL11M12A01-02E, and added ZR40H Automatic Calibration Unit to CMPL11M12A01-11E
Oct. 2000/1st Edition
Newly published
Yokogawa Electric Corporation
2-9-32 Nakacho, Musashino-shi, Tokyo 180-8750, JAPAN
Homepage: http://www.yokogawa.com/
IM 11M12A01-02E
12th Edition : Apr. 27, 2017-00
Blank Page