Yokogawa Model ISC40G(S) Sensors and Fittings for Inductive Conductivity Measurement Manual
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
User’s
Manual
Model ISC40G(S)
Sensors and Fittings for Inductive
Conductivity Measurement
IM 12D08J02-E-E
18th edition
IM 12D08J02-E-E_ed18.indd 1
03/03/16 12:58
2
(BG)
Всички улътвания за продукти от серията
ATEX Ex ce лредлагат на английски
език. Ако се нуждаете oт улътвания за
продукти oт серията Ex на родния ви
eзик, ce свържете c най-близкия oфис или
представителство на фирма Yokogawa.
(CZ)
Všechny uživatelské příručky pro výrobky, na
něž se vztahuje nevýbušné schváleni ATEX
Ex, jsou dostupné v angličtině. Požadujete-li
pokyny týkající se výrobků s nevýbušným
schválením ve vašem lokálním jazyku,
kontaktujte prosím vaši nejbližší reprezentačni
kancelář Yokogawa.
(D)
Alle Betriebsanleitungen für ATEX Ex
bezogene Produkte stehen in den Sprachen
Englisch. Sollten Sie die Betriebs- anleitungen
für Ex-Produkte in Ihrer Landessprache
benötigen, setzen Sie sich bitte mit Ihrem
örtlichem Yokogawa-Vertreter in Verbindung.
(E)
Todos los manuales de instrucciones para
los productos antiexplosivos de ATEX están
disponibles en inglés. Si desea solicitar las
instrucciones de estos artículos antiexplosivos
en su idioma local, deberá ponerse en contacto
con la oficina o el representante de Yokogawa
más cercano.
(F)
Tous les manuels d’instruction des produits
ATEX Ex sont disponibles en langue anglaise.
Si vous nécessitez des instructions relatives
aux produits Ex dans votre langue, veuillez
bien contacter votre représentant Yokogawa le
plus proche.
(GB)
All instruction manuals for ATEX Ex related
products are available in English. Should you
require Ex related instructions in your local
language, you are to contact your nearest
Yokogawa office or representative.
(DK)
Alle brugervejledninger for produkter relateret
til CE er tilgængelige på engelsk. Skulle De
ønske yderligere oplysninger om håndtering
af CE produkter på eget sprog, kan De rette
henvendelse herom til den nærmeste Yokogawa
afdeling eller forhandler.
(GR)
Ολα τα εγχειριδια λειτουργιας των
προιοντων με ATEX Ex διατιθενται στα
Αγγλικα. Σε περιπτωση πον χρειαζεοτε
οδηγιεξ σχετικα με Ex στην τοπικη
γλωσσα παρακαλονμε επικοινωνηστε με
το πλησιεστερο γραφειο τηζ Yokogawa η
αντιπροσωπο τηξ.
(EST)
Kõik ATEX Ex toodete kasutamisjuhendid on
esitatud inglise keeles. Ex seadmete muukeelse
dokumentatsiooni saamiseks pöörduge lähima
lokagava (Yokogawa) kontori või esindaja
poole.
(H)
Az ATEX Ex mûszerek gépkönyveit angol
nyelven adjuk ki. Amennyiben helyi nyelven
kérik az Ex eszközök leírásait, kérjük keressék
fel a legközelebbi Yokogawa irodát, vagy
képviseletet.
(I)
Tutti i manuali operativi di prodotti ATEX
contrassegnati con Ex sono disponibili in
inglese. Se si desidera ricevere i manuali
operativi di prodotti Ex in lingua locale,
mettersi in contatto con l’ufficio Yokogawa più
vicino o con un rappresentante.
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 2
03/03/16 12:58
3
(LV)
Visas ATEX Ex kategorijas izstrâdâjumu
Lietoðanas instrukcijas tiek piegâdâtas
angïu valodâs. Ja vçlaties saòemt Ex ierîèu
dokumentâciju citâ valodâ, Jums ir jâsazinâs ar
firmas Jokogava (Yokogawa) tuvâko ofisu vai
pârstâvi.
(LT)
Visos gaminiø ATEX Ex kategorijos
Eksploatavimo instrukcijos teikiami anglø
kalbomis. Norëdami gauti priestaisø Ex
dokumentacijà kitomis kalbomis susisiekite su
artimiausiu bendrovës Yokogawa biuru arba
atstovu.
(PL)
Wszystkie instrukcje obsługi dla urządzeń
w wykonaniu przeciwwybuchowym Ex,
zgodnych z wymaganiami ATEX, dostępne
są w języku angielskim. Jeżeli wymagana
jest instrukcja obsługi w Państwa lokalnym ję
zyku, prosimy o kontakt z najbliższym biurem
Yokogawy.
(RO)
Toate manualele de instructiuni pentru
produsele ATEX Ex sunt in limba engleza.
In cazul in care doriti instructiunile in limba
locala, trebuie sa contactati cel mai apropiat
birou sau reprezentant Yokogawa.
(M)
Il-manwali kollha ta’ l-istruzzjonijiet għal
prodotti marbuta ma’ ATEX Ex huma
disponibbli bl-Ingliż. Jekk tkun teħtieģ
struzzjonijiet marbuta ma’ Ex fil-lingwa
lokali tiegħek, għandek tikkuntattja lill-eqreb
rappreżentan jew uffiććju ta’ Yokogawa.
(S)
Alla instruktionsböcker för ATEX Ex
(explosionssäkra) produkter är tillgängliga
på engelska. Om Ni behöver instruktioner för
dessa explosionssäkra produkter på annat språk,
skall Ni kontakta närmaste Yokogawakontor
eller representant.
(NL)
Alle handleidingen voor producten die te
maken hebben met ATEX explosiebeveiliging
(Ex) zijn verkrijgbaar in het Engels. Neem,
indien u aanwijzingen op het gebied van
explosiebeveiliging nodig hebt in uw eigen taal,
contact op met de dichtstbijzijnde vestiging van
Yokogawa of met een vertegenwoordiger.
(SF)
Kaikkien ATEX Ex-tyyppisten tuotteiden
käyttöhjeet ovat saatavilla englannin-. Mikäli
tarvitsette Ex-tyyppisten tuotteiden ohjeita
omalla paikallisella kielellännne, ottakaa
yhteyttä lähimpään Yokogawa-toimistoon tai
-edustajaan.
(P)
Todos os manuals de instruções referentes aos
produtos Ex da ATEX estão disponíveis em
Inglês. Se necessitar de instruções na sua língua
relacionadas com produtos Ex, deverá entrar
em contacto com a delegação mais próxima ou
com um representante da Yokogawa.
(SK)
Všetky návody na obsluhu pre prístroje s
ATEX Ex sú k dispozícii v jazyku anglickom.
V prípade potreby návodu pre Ex-prístroje vo
Vašom národnom jazyku, skontaktujte prosím
miestnu kanceláriu firmy Yokogawa.
(SLO)
Vsi predpisi in navodila za AEX Ex sorodni
pridelki so pri roki v anglišèini. Èe so Ex
sorodna navodila potrebna v vašem tukejnjem
jeziku, kontaktirajte vaš najbliši Yokogawa
office ili predstaunika.
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 3
03/03/16 12:58
4
Table of Contents
1 INTRODUCTION
7
2
GENERAL SPECIFICATIONS
9
3
INSTALLATION OF SENSOR WITH STANDARD OPTIONS 14
Typical installation
3.1.1 Installation of ISC40G(S)-GG (TG) with flange adapters
3.1.2 Installation of ISC40G(S)-TG with T-piece
3.1.3 Installation of ISC40G(S)-GS with flange adapters
14
14
16
17
4
DIMENSIONS ISC40 SENSOR
18
5
WIRING ISC40 SENSOR
19
6
MODELCODE ISC40 SENSOR
20
7
SPAREPARTS ISC40 SENSOR
21
8
EU DECLARATION OF CONFORMITY
22
9
ISC40FS – FLOW FITTING SUBASSEMBLY
23
1.1 General
1.2 Unpacking and Checking
1.3 Warranty and Service
1.4 Serial Number definition
2.1 Measuring elements
2.2 Materials
2.3 Functional specifications (at 25°C)
2.4 Dynamic specifications
2.5 Operating range
2.6 Regulatory standards
2.7 Shipping details
2.8 Environmental conditions
2.9 Process connections
3.1
9.1
9.2
9.3
9.4
9.5
General Specifications
9.1.1 Materials
9.1.2 Operating range
9.1.3 Shipping details
9.1.4 Process connections
Installation of ISC40 sensor in ISC40FS
Dimensions ISC40FS
Modelcode ISC40FS
Spareparts ISC40FS
7
7
7
8
9
9
9
9
10
10
11
11
11
23
23
23
23
23
23
24
25
25
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 4
03/03/16 12:58
5
10 ISC40FF – FLOW FITTING
26
11 ISC40FD – IMMERSION FITTING
30
12
34
10.1
10.2
10.3
10.4
10.5
11.1
11.2
11.3
11.4
11.5
General Specifications
10.1.1 Materials
10.1.2 Operating range
10.1.3 Shipping details
10.1.4 Process connections
Installation of ISC40 sensor in ISC40FF
Dimensions ISC40FF
Modelcode ISC40FF
Spareparts ISC40FF
General Specifications
11.1.1 Materials
11.1.2 Operating range
11.1.3 Process connections
Installation of ISC40 sensor in ISC40FD
Dimensions ISC40FD
Modelcode ISC40FD
Spareparts ISC40FD
Chemical Compatibility Chart
26
26
26
26
26
26
27
28
29
30
30
30
30
30
32
33
33
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 5
03/03/16 12:58
6
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 6
03/03/16 12:58
7
1 INTRODUCTION
1.1 General
The sensor and fitting program for inductive
conductivity measurement (model ISC40)
is designed to meet the most common
installation requirements in terms of material
compatibility, process connections and
flow dynamics. The various installation
possibilities are described and illustrated in
this manual.
The following categories of installation can
be recognised:
1. Direct mounting of sensors in tank wall or
customer supplied flanges
2. Cost effective installation of sensors
using process adapters
3. Maintenance friendly installation of
sensors using in-line subassemblies
4. Installation in flow chambers for
measurements in sample streams or
bypass loops
5. Installation in immersion fittings for
measurement in open tanks or channels
Within the wide range of process adapters,
subassemblies, flow fittings and immersion
fittings it is easy to find the appropriate
installation that fits the application (a wide
choice of materials), the plant installation
practice (a wide choice of process connections) and the maintenance procedures.
1.2 Unpacking and Checking
Upon delivery, unpack the sensor carefully
and inspect it to ensure that it is not
damaged during shipment. If damage is
found, retain the original packing material
and immediately notify the carrier and the
relevant local Yokogawa Sales office.
Make sure the Model Code and Serial
Number on the sensor are the same as on
the packing list. Also check if option(s) that
were ordered, are included and correct.
1.3 Warranty and Service
Yokogawa products are guaranteed free
from defects in workmanship and materials
under normal use and service for a period
of (typically) 12 months from the date of
shipment from the manufacturer. Individual
Sales organizations can deviate from the
typical warranty period, and the conditions
of sale relating to the original purchase
order should be consulted. Damage caused
by wear and tear, inadequate maintenance,
corrosion, or by the effects of chemical
processes is excluded from this warranty
coverage. In the event of a warranty claim,
the defective goods should be sent (freight
paid) to the Service Department of the
relevant Yokogawa Sales office for repair or
replacement (at Yokogawa’s discretion).
The following information must be included
in the letter accompanying the returned
goods:
• Model Code and Serial Number.
• Original Purchase Order and Date.
• Length of time in service and description
of the process.
• Description of the fault and circumstances
of the failure.
• Process/environmental conditions that
may be related to the failure of the sensor.
• Statement as to whether warranty or nonwarranty service is requested.
• Complete shipping and billing instructions
for return of material, plus the name and
phone number of a contact person that
can be reached for further information.
• Clean Statement
Returned goods that have been in
contact with process fluids must be
decontaminated and disinfected prior
to shipment. Goods should carry a
certificate to this effect, for the health and
safety of our employees. Material Safety
Data sheets must be included for all
components of the process to which the
sensor(options) have been exposed.
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 7
03/03/16 12:58
8
1.4 Serial Number definition
The Serial Number is defined by nine (9)
alphanumeric characters:
X1X2
Production Location
X3X4
Year/Month code
X5X6X7X8X9 Tracking number
Example: N3P600028
Method used for year/month numbering
Table 1: Production Year code
Table 2: Production Month code
Year Year code
2014 P
2015 R
2016 S
2017 T
2018 U
2019 V
2020 W
2021 X
2022 Y
2023 Z
2024 1
2025 2
Month
January
February
March
April
May
June
July
August
September
October
November
December
Year
Year code
2026 3
2027 4
2028 5
2029 6
2030 7
2031 8
2032 9
2033 A
2034 B
2035 C
2036 D
2037 E
Month code
1
2
3
4
5
6
7
8
9
A
B
C
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 8
03/03/16 12:58
9
2 GENERAL SPECIFICATIONS
The ISC40 inductive conductivity sensor is suitable for use with the Yokogawa inductive
conductivity analyzers.
2.1 Measuring elements
:T
oroïds with high permeability magnetic material
Pt1000 or 30k temperature element
2.2 Materials
Wetted parts sensor
Body ISC40*-G*
: 30% glass filled PEEK, FDA approved
Body ISC40*-T*
: PFA, FDA approved, PIM regulation 10/2011 approved
Non-wetted parts sensor
Sealing gasket
: Viton
Thread part
: AISI 316 SS
Options for sensor
All options except /TFD : AISI 316 SS and O-ring material as wetted part
/TFD
: AISI 316 SS as non-wetted part
TFM and Kalrez as wetted part
2.3 Functional specifications (at 25°C)
Temperature element : Pt1000 to IEC 751
30k NTC
Installation factor
: 1.88 cm-1 nominal for PEEK sensor
3.00 cm-1 nominal for PFA sensor
Actual installation can change this factor.
If there is less than 25mm spacing between sensor and holder,
in-situ calibration is necessary to meet the specified accuracies
(see fig. 1)
Only for PEEK
Correction factor (x nominal C.C.)
1.30
1.25
1.20
1.15
non conductive
piping
1.10
D
1,05
1.00
0.95
conductive
piping
0.90
0
10
D in millimeters
20
30
40
50
Fig 1: A
ctual installation factor as function of spacing around the sensor
Note: T
he ISC40 temperature sensor is designed for cell compensation and for indication.
It is NOT designed for process temperature control.
2.4 Dynamic specifications
Response time
: t90 < 5 min. for PEEK sensor
conductivity
: t90 < 10 min. for PFA sensor
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 9
03/03/16 12:58
10
2.5 Operating range
Conductivity
: 0 – 2000 mS/cm at actual process temperature.
Note: The sensor has an error (0.5 µS/cm for PEEK model, 1.0 µS/cm for PFA model) that
must be considered when application is chosen.
Temperature
Pressure
Cable length
: - 20°C to 130°C (-4°F to 266°F)
: 0 to 20 bar (0 to 290 PSIG) for PEEK sensor
0 to 15 bar (0 to 217 PSIG) for PFA sensor.
: max 50 meter, in combination with WF10 extension cable and
BA10 junction box
2.6 Regulatory standards
CE
: Decision 768/2008/EC
- ATEX
: Directive 94/9/EC, as amended by Regulation (EC) no.
1882/2003
Certificate no.
: DEKRA 11ATEX0063 X
II 1 G Ex ia IIC T4…T6 Ga
Electrical data
: Sensor output circuits (permanently connected cable)
connected to a
certified intrinsically safe circuit with the following maximum
values: Ui = 19.1 V; Ii = 170 mA; Pi = 0.8 W; Ci = 0 nF; Li = 0 mH
or
certified intrinsically safe Yokogawa Inductive Conductivity
transmitter model FLXA21 series, model ISC202S series or
model IC200S series
Note: T
he effective internal capacitance Ci and the effective internal inductance Li of the
sensor are depending only upon the properties and the length of the connected cable.
Special conditions (X) : T
6 for Tamb. -30°C to 40°C
T5 for Tamb. -30°C to 95°C
T4 for Tamb. -30°C to 130°C, depending on sensor body
material
: The sensor must be installed and used so, that dangers
of ignition due to hazardous electrostatic charges cannot
occur, especially in the case that the process medium is nonconductive.
- Pressure
: Directive 97/23/EC, as amended by Regulation (EC) no.
1882/2003
Applying article
: 3.3 (Sound Engineering Practice)
IECEx
Applying standards
Certificate no.
: IEC 60079-0 : 2007
IEC 60079-11 : 2006
IEC 60079-26: 2006
: IECEx DEK 11.0028X
Ex ia IIC T4…T6 Ga
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 10
03/03/16 12:58
11
CSA
Certificate no.
:2
447837 IS, Class I Div. 1, GP A, B, C, D T4…T6
Master Contract no 182892
Electrical data
: Sensor output circuits (permanently connected cable)
connected to a certified intrinsically safe circuit with the
following maximum values: Uo = 19.1 V; Io = 170 mA;
Po = 0.8 W; Co≥ Ci + Ccable; Lo≥ Li + Lcable
or
certified intrinsically safe Yokogawa Inductive Conductivity
transmitter model FLXA21 series, model ISC202S series or
model IC200S series
Ambient
: T6 for Tamb. -30°C to 40°C
temperature range
T
5 for Tamb. -30°C to 95°C
T4 for Tamb. -30°C to 130°C, depending on sensor body
material
Note: Intrinsically safe when connected as per Control Drawing FF1-K1244QY (see fig. 2)
FM
Certificate no.
Electrical data
:3
046320 IS, Class I, Div. 1, Groups A, B, C, D T4…T6
: Sensor output circuits (permanently connected cable)
connected to a FM approved intrinsically safe apparatus
meeting the entity parameters of the ISC40S:
Uo≤ 19.1 V; Io≤ 170 mA; Po≤ 0.8 W;
Co≥ Ci + Ccable; Lo≥ Li + Lcable
or
certified intrinsically safe Yokogawa Inductive Conductivity
transmitter model FLXA21 series, model ISC202S series or
model IC200S series
Ambient
: T6 for Tamb. -30°C to 40°C
temperature range
T
5 for Tamb. -30°C to 85°C
T4 for Tamb. -30°C to 85°C, depending on sensor body
material
Note: Intrinsically safe when connected as per Control Drawing FF1-K1244QX (see fig. 3)
2.7 Shipping details
Package size (LxWxH)
ISC40*-**-*-03 (05) : 350 x 270 x 50 mm (13.8 x10.6 x 2.0 inch)
ISC40*-**-*-10 (15, 20) : 320 x 240 x110 mm (12.6 x 9.5 x 4.3 inch)
Package weight (app.)
ISC40*-**-*-03
: 1.0 kg (2.2 lbs)
ISC40*-**-*-05
: 1.3 kg (2.9 lbs)
ISC40*-**-*-10
: 1.6 kg (3.5 lbs)
ISC40*-**-*-15
: 2.1 kg (4.6 lbs)
ISC40*-**-*-20
: 2.5 kg (5.5 lbs)
2.8 Environmental conditions
Storage temperature : -30°C to 50°C (-22°F to 122°F)
Water proof
: IP67 (conform IEC 60529), also in combination with the
preferred Yokogawa process connections
2.9 Process connections
Process connections are made in combination with a variety of adapters and fittings,
which are available in AISI 316 SS, PVC or PVDF (see relevant sections in this manual).
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 11
03/03/16 12:58
12
Control Drawing CSA
The ISC40S sensor shall be installed with
one of the Yokogawa transmitters model:
• ISC202S • IC200S • FLXA21
with following parameters:
Uo
Io
Po
Lo
Co
ISC202S IC200S
14.4 V
19.1 V
88 mA 162 mA
317 mW
4.5 mH 800µH
600 nF
254nF
FLXA21
11.76 V
60.6 mA
178 mW
8 mH
100 nF
ISC40 sensor
terminals 11-17.
Maximum cable
length: 50m.
or
To a CSA approved intrinsically safe
apparatus meeting the entity parameters of
the ISC40S: Uo ≤ 19.1V
Io ≤ 170mA
Po ≤ 0.8W
Co ≥ Ci + C(cable)
Lo ≥ Li + L(cable)
The effective inductive capacitance Ci and the
effective induced inductance Li of the sensor
depends only upon the properties and the
length of the connected cable (max 50m).
When installing this equipment, follow the
manufacturer’s control drawing. Installing
should be in accordance with Canadian
Electrical Code Part 1 or CEC Part1.
Fig 2: F
F1-K1244QY
Control Drawing CSA
Warning: To prevent ignition of flammable or combustible atmospheres, disconnect power
before servicing or read, understand and adhere to the manufacturer’s live maintenance
procedures.
ISC40
Cable length
in meters, any number between 1 and 50 meters
Temperature Element
T1 PT1000
T3 30k thermistor
Plastic and adaption code
GG Glass filled PEEK, general model
GR Glass filled PEEK, retractable model
GS Glass filled PEEK, shaft model
TG PFA, general model
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 12
03/03/16 12:58
13
Control Drawing FM
The ISC40S sensor shall be installed with
one of the Yokogawa transmitters model:
• ISC202S • IC200S • FLXA21
with following parameters:
ISC202S IC200S FLXA21
Uo
14.4 V
19.1 V
11.76 V
Io
88 mA 162 mA 60.6 mA
Po
317 mW
178 mW
Lo
4.5 mH 800µH
8 mH
Co
600 nF
254nF
100 nF
or
To a FM approved intrinsically safe
apparatus meeting the entity parameters of
the ISC40S: Uo ≤ 19.1V
Io ≤ 170mA
Po ≤ 0.8W
Co ≥ Ci + C(cable)
Lo ≥ Li + L(cable)
The effective inductive capacitance Ci and the
effective induced inductance Li of the sensor
depends only upon the properties and the
length of the connected cable (max 50m).
When installing this equipment, follow the
manufacturer’s control drawing. Installing
should be in accordance with ANSI/ISA
RP 12.06.01 “Installation of Intrisically
Safe Systems for Hazardous (Classified)
Locations” and the National Electrical Code
(ANSI/NFPA 70).
Ta 85ºC/85ºC/40ºC
Fig 3: FF1-K1244QX
Control Drawing FM
Warning: To prevent ignition of flammable or combustible atmospheres, disconnect power
before servicing or read, understand and adhere to the manufacturer’s live maintenance
procedures.
ISC40
Cable length
in meters, any number between 1 and 50 meters
Temperature Element
T1 PT1000
T3 30k thermistor
Plastic and adaption code
GG Glass filled PEEK, general model
GR Glass filled PEEK, retractable model
GS Glass filled PEEK, shaft model
TG PFA, general model
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 13
03/03/16 12:58
14
3 INSTALLATION OF SENSOR WITH STANDARD OPTIONS
For optimum measurement results,
the ISC40 sensor should be installed
in a location that offers an acceptable
representation of the process composition
and DOES NOT exceed the specifications
of the sensor.
A
It is important that the process flow is directed
through the hole in the donut of the sensor.
For this reason the flats on the sensor top
part have to be installed perpendicular on the
process flow (see fig. 4).
The inductive conductivity measurement
technique requires a process fluid
surrounding the donut of the sensor. The
installation factor mentioned on the label
of the sensor cable will assure accurate
conductivity measurement under the
condition that the donut is surrounding by
25 mm (1”) process fluid. If this condition
cannot be met, the measurement loop
(analyser in combination with sensor) has
to be calibrated. Consult the instruction
manuals of the analyser for details.
The mounting of the sensor in the process
adapters is described in section 3.1. The
mounting of the process adapters in the
end-users application requires compatibility
of materials and process connections.
3.1
B
Fig 4: A
ligning of sensor donut (B) and
flats (A)
1
2
PUSH & TURN
Typical installation
3.1.1 Installation of ISC40G(S)-GG (TG)
with flange adapters
The ISC40G(S)-GG (TG) sensor is
supplied with a flat Viton sealing gasket
for compatibility with existing installations.
When using the sensor in combination with
the new fittings, which are improved by the
addition of O-ring seals (see figure 6), the
flat Viton gasket should be discarded.
It is important that the access port has a
diameter of at least 50 mm (2”) to allow
insertion of the donut shaped end of the
sensor. For more detailed dimensions of the
sensor see paragraph 4.
3
TORQUE MAX. 25Nm.
DO NOT WRENCH
Fig 5: M
ounting Procedure ISC40G(S)GG (TG) with flange adapters
Note: When a torque of more than 25 Nm
is applied, the top part of the body of the
PFA sensor will be damaged (TG model)
Note: Retighten after 24 hours
IM 12D8J2-E-E
IM 12D08J02-E-E_ed18.indd 14
03/03/16 12:58
15
The sensor cable and mounting thread
are pulled through the hole of the flange,
and the sensor is sealed from the process
by tightening the mounting nut. Turning
of the sensor by the torque forces can be
avoided by using a wrench on the flats
on top of the sensor (see fig. 5). Sensor
is installed correctly if the flats are aligned
perpendicular on the process flow.
Flange dimensions in mm (inches)
O-ring
FLANGE ADAPTER
d
26.65 x 2.62
O-ring
40.87 x 3.53
26.57 x 3.53
116 (4.56)
40.64 x 5.33
D1
D2
Fig 6: Option /SFA, /SFD,
/S2W, /STW
Option
d D1 D2
/SFA
Ø19 (0.75)
121 (4.76)
152 (6.0)
/SFD
Ø18 (0.71)
125 (4.92)
165 (6.5)
/TFD
Ø18 (0.71)
145 (5.71)
185 (7.3)
OK
OK
OK
OK
Not recommended
(solid deposits)
Not recommended
(air entrapment)
Fig 7: Installation examples
IM 12D8J2-E-E
IM 12D08J02-E-E_ed18.indd 15
03/03/16 12:58
16
3.1.2 Installation of ISC40G(S)-TG
with T-piece
The ISC40G(S)-TG sensor is a PFA sensor
which can be used in combination with a
T-piece of which the innerside is lined with
PFA. This can be done using option /TFD
(sealing material with stainless steel DN65
PN10 flange), or by option /TFN (sealing
material) if the stainless steel flange DN65
PN10 is already available. Both options are
available as sparepart.
DN65 PN10 flange
TFM ring
O-ring 29.74 x 3.53
Goretex seal
123 (4.83”)
Note: If sensor is replaced, you have to
use new sealing materials (part
no. K1541XG) to prevent process
leakage.
8 (0.31”)
SS304 Nut
ø100 (3.94”)
PN10 + TFM ring = TFD
TFM ring alone = TFN
ISC40.-TG/TFD
Fig 8: Flange adapters
(option /TFD, /TFN) for –TG sensor
in combination with T-piece
T-piece:
Partno. Flanges
K1500HG
DN80 PN10
K1500HF
DN100 PN10
Description
T-piece, DN80 flange
T-piece, DN100 flange
C
Mounting flange and bolts are included
øB
øA
DN65 PN10
DN80 PN10
DN100 PN10
A
185
200
220
B
C
145 4x ø18
160 8x ø18
180 8x ø18
100.0
DN80 PN10 / DN100 PN10
DN65 PN10
300.0
Innerside of T-Piece is lined with PFA
Thickness: 4mm. Flange material: SS316
Fig 9: T-piece dimensions
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 16
03/03/16 12:58
17
3.1.3 Installation of ISC40G(S)-GS
with flange adapters
plastic nut
The ISC40G(S)-GS sensor is designed for
sanitary applications. For these applications
special process connections are necessary.
tube
Mounting procedure:
• Screw the tube completely in the
stainless steel nut.
• Thread the sensor cable through
the flange adapter parts in the right
sequence.
• Screw the tube hand tight into the flange,
a mechanical stop will be felt.
• Tighten the plastic nut onto the sensor;
screw the plastic nut completely tight.
• Tighten the stainless steel nut for locking
the sensor.
Sensor is installed correctly if the flats on
the sensor are aligned perpendicular on the
process flow.
nut
ring
O-ring 18.7 x 2.6
Tuchenhagen
Flange
Check the O-ring
ø 60 x 3 for correct
mounting.
ISC40G(S)-GS
Fig 10: Installation /SFT
Sanitary 2” Tri-clamp
Tuchenhage
Sanitary 2” Tri-clamp complete
O-ring
18.72 x 2.62
123 (4.84")
O-ring
18.72 x 2.62
134,3 (5.30")
O-ring
18.72 x 2.62
2” clamp steal
Ø71 (2.80")
/SFT
57,3 (2.25")
72 (2.83")
19 (0.75")
/STC1
/STC2
Fig 11: Option /SFT, /STC1, /STC2
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 17
03/03/16 12:58
18
4 DIMENSIONS ISC40 SENSOR
Unit: mm (inches)
ø17 (0.67”)
nut G3/4
viton gasket
For mounting in
ISC40PR retractable
fitting
124 (4.88")
Minimum hole
to fit sensor
through:
GG Ø48 (1.89")
TG Ø50 (1.99”)
182 (7.16")
Ø40 (1.57")
Minimum hole
to fit sensor
through: Ø48 (1.89")
m19 x 1.5
55 (2.16")
Ø47 (1.85")
GG Ø47 (1.85”)
TG Ø49 (1.93”)
Fig 12: ISC40G(S)-GG (TG)
Fig 13: ISC40G(S)-GR
ISC40-PR
For mounting
in SFT, STC 1 and
STC 2 Fittings.
Ø20 (0.78")
193 (7.60")
Stainless steel
77 (3.0")
Peek
Ø47 (1.85")
Minimum hole
to fit sensor
through: Ø48 (1.89")
Fig 14: ISC40G(S)-GS
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 18
03/03/16 12:58
19
5 WIRING ISC40 SENSOR
The ISC40 sensor is provided with a fixed cable. The connections of this dual coax cable
and the Yokogawa Inductive Conductivity analyser are given in table 1.
Table 1: Definition cable and analyzer
Cable wire color
ISC analyser terminal #
Red 11
Blue 12
Yellow 14
White
15
White (shield)
16
Brown
13
Brown (shield)
17
Signal description
Temperature
Temperature
Shield
Primary coil
Primary coil
Secondary coil
Secondary coil
ISC40 sensor
terminals 11-17.
Maximum cable
length: 50m.
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 19
03/03/16 12:58
20
6 MODELCODE ISC40 SENSOR
Model Code Suffix
Option
ISC40G
Sensor
-GG
type
-GR
-GS
-TG
Temperature
-T1
sensor
-T3
Cable length
-3
-5
-10
-15
-20
Options for Sensor
Flange adapters -GG, -TG /SFA
/SFD
/STW
/S2W
/TFD
/TFN
Flange adapters for -GS /SFT
/STC1
/STC2
Protection Hose
for-TF, -TG, -GG /PH
Certificates
/M
/Q
Description
General purpose inductive conductivity sensor
Glass filled PEEK, general model
Glass filled PEEK, retractable model
Glass filled PEEK, shaft model
PFA, general model
Pt1000
30k thermistor,for IC200 select only T3
03 meter
05 meter
10 meter
15 meter
20 meter
Material
Proc.Connection
AISI 316 SS
2” ANSI 150 lbs
AISI 316 SS
NW50
AISI 316 SS
3” tri-clamp
AISI 316 SS
2” tri-clamp
TFM, AISI 316 SS NW65-PN10
TFM
For NW65-PN10
AISI 316 SS
Sanitary Tuchenhagen
AISI 316 SS
Sanitary 2” tri clamp
AISI 316 SS
Tri-clamp complete
03m /05m /10m /15m /20m
Same length as the cable
Material certificate Only for flange adapters
Quality certificate except /TFD and /TFN
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 20
03/03/16 12:58
21
7 SPAREPARTS ISC40 SENSOR
Part no. Description
K1500AM Gasket
K1500AL Mounting nut
Parts ISC40 sensor
Material
Quantity
Viton
5
AISI 316 SS
3
Options ISC40 sensor, Flange adapters
Part no. Description Process connection
Material
O-ring(s)
K1541ZR /SFA
2” ANSI 150 lbs
AISI 316 SS
Viton
K1541ZQ /SFD
NW50
AISI 316 SS
Viton
K1541KB /STW
3” tri-clamp
AISI 316 SS
EPDM
K1541KC /S2W
2” tri-clamp
AISI 316 SS
EPDM
K1541XF /TFD
DN65 PN10
AISI 316 SS,TFM Kalrez
K1541XG /TFN
used with DN65 PN10
TFM
Kalrez
K1541ZP /SFT
Sanitary Tuchenhagen
AISI 316 SS
EPDM
K1541ZG /STC1
Sanitary 2” tri-clamp
AISI 316 SS
EPDM
K1541ZF /STC2
Tri-clamp complete
AISI 316 SS
EPDM
K1500HG
T-piece, DN80 flange
DN80 PN10
K1500HF
T-piece, DN100 flange
DN100 PN10
Note: Other O-ring materials are available as sparepart
Part no.
K1500CA
K1500CB
K1500CC
K1500CD
K1500CH
K1541ZK
K1541ZH
K1500DJ
K1500DK
K1500AH
K1500CM
K1500CQ
K1500CP
K1500CR
K1500CT
K1500CS
K1500CU
O-rings ISC40 sensor, Flange adapters
Description Dimensions
Material
O-rings /SFA, /SFD
O-ring set 40.64 x 5.33; 26.57 x 3.53
EPDM
O-ring set 40.64 x 5.33; 26.57 x 3.53
Viton
O-ring set 40.64 x 5.33; 26.57 x 3.53
Silicon
O-ring
40.64 x 5.33
Kalrez
O-ring
26.57 x 3.53
Kalrez
O-rings /STW
O-ring set 40.87 x 3.53; 26.65 x 2.62; 3” seal-clamp EPDM
O-rings /S2W
O-ring set 40.87 x 3.53; 26.65 x 2.62; 2” seal-clamp EPDM
O-ring set 40.87 x 3.53; 26.65 x 2.62; 2” seal-clamp Viton
O-ring set 40.87 x 3.53; 26.65 x 2.62; 2” seal-clamp Silicon
O-rings /TFD, /TFN
O-ring
29.74 x 3.53
Kalrez
O-rings /SFT
O-ring set 18.72 x 2.62; 60 x 3
EPDM
O-rings /STC1
O-ring
18.72 x 2.62
EPDM
O-ring
18.72 x 2.62
Viton
O-ring
18.72 x 2.62
Silicon
O-rings /STC2
O-ring set 18.72 x 2.72; 2” seal-clamp
EPDM
O-ring set 18.72 x 2.72; 2” seal-clamp
Viton
O-ring set 18.72 x 2.72; 2” seal-clamp
Silicon
Quantity
5 sets
5 sets
5 sets
1
1
2 sets
2 sets
2 sets
2 sets
1
5 sets
5
5
5
5 sets
5 sets
5 sets
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 21
03/03/16 12:58
22
8 EU DECLARATION OF CONFORMITY
YOKOGAWA
EU DECLARATION OF CONFORMITY
We:
Yokogawa Process Analyzers Europe B.V.
Euroweg 2
3825 HD Amersfoort
The Netherlands
herewith declare under our sole responsibility that the products, model: ISC40G and ISC40S
further specified with model suffix- and option codes: As listed in Annex-1 in this document
are manufactured in accordance with the requirements for CE-marking of products as stated in EC Decision:
768/2008/EC on a common framework for the marketing of products
by applying the following standards:
EN-ISO 9001: 2008
Quality management systems - Requirements
Model ISC40G and ISC40S are:
• In compliance with the essential requirements of the specific product legislation:
- Pressure Equipment
Directive 97/23/EC (PED)
As amended by Regulation (EC) no. 1882/2003, by applying:
Article 3.3:
Sound Engineering Practice
- RoHS
by applying:
Category 9:
•
Directive 2011/65/EU
Industrial monitoring and control instruments
Produced according to appropriate quality control procedures.
Model ISC40S is:
• In compliance with the essential requirements of the specific product legislation:
- Explosive atmospheres
Directive 94/9/EC (ATEX)
As amended by Regulation (EC) no. 1882/2003, by applying the following standards:
EN 60079-0: 2009
Explosive atmospheres – Part 0: Equipment – General requirements
EN 60079-11: 2007
Explosive atmospheres – Part 11: Equipment protection by intrinsic safety “i”
EN 60079-26: 2007
Explosive atmospheres – Part 26: Equipment with equipment protection
level (EPL) Ga
The provisions fulfilled are: II 1 G Ex ia IIC T4...T6 Ga
Number of the EC-type Examination Certificate: DEKRA 11 ATEX 0063 X
Name of the notified body: DEKRA Certification B.V.
Identification number of the notified body: 0344
Address of the notified body: Meander 1051, 6825 MJ Arnhem, The Netherlands
The CE-mark has been affixed on the product in 2000 for the first time.
If applicable, the product is checked against the latest official released revision of the standards mentioned
above; differences do not affect the certified product identified on this declaration.
Amersfoort - July 01, 2014
H. Leijten
General Manager
Yokogawa Process Analyzers Europe B.V.
1/2
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 22
03/03/16 12:58
23
9
ISC40FS – FLOW FITTING SUBASSEMBLY
9.1 General Specifications
9.1.1 Materials
Wetted parts
Model ISC40FS-FCSA
Model ISC40FS-PCSA
Model ISC40FS-SCSA
Model ISC40FS-SCWN
:
:
:
:
PVDF (Kynar); Viton
Polypropylene; Viton
AISI 316 SS; Viton
AISI 316 SS; Viton
:
:
:
:
Max.
Max.
Max.
Max.
Non-wetted parts
Nut : AISI 304 SS
9.1.2 Operating range
Temperature
Model ISC40FS-FCSA
Model ISC40FS-PCSA
Model ISC40FS-SCSA
Model ISC40FS-SCWN
Pressure
Model ISC40FS-FCSA
130°C
100°C
150°C
150°C
(266°F)
(212°F)
(302°F)
(302°F)
Model ISC40FS-SCSA
Model ISC40FS-SCWN
: Max.
Max.
: Max.
Max.
: Max.
: Max.
9.1.3 Shipping details
Package size (LxWxH)
Package weight (max.)
: 215 x 150 x 55 mm (8.46 x 5.90 x 2.17 inch)
: 0.9 kg (2.0 lbs)
Model ISC40FS-PCSA
10 bar (150 PSI) at 20°C (68°F)
1 bar (15 PSI) at 130°C (266°F)
6 bar (90 PSI) at 20°C (68°F)
1 bar (15 PSI) at 100°C (212°F)
10 bar (150 PSI) at operating temperature
10 bar (150 PSI) at operating temperature
9.1.4 Process connections
2” screw-in coupling or weld-in coupling
Note: T
he ISC40 sensor type GG and TG are supplied with a Viton gasket. This gasket may
be used with the older series of ISC40FS fittings. It should be discarded when the
sensor is mounted in combination with the ISC40FS type FCSA and PCSA.
9.2 Installation of ISC40 sensor in ISC40FS
The key difference between a sensor adapter and a subassembly is, that with subassembly
installation the sensor can be removed from the process installation without removing
the subassembly first. This allows easier access to the sensor for maintenance activities.
Generally, the subassembly consists of three parts of which one part is fitted permanently
to the process installation (welded or threaded); the second part is fitted to the sensor and
the third part holds the earlier parts together. In addition to these parts there are elastomeric
seals were appropriate. Standard O-ring material is Viton, other O-ring materials are
available as sparepart (see Figs 15 and 16).
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 23
03/03/16 12:58
24
O-ring
40.64 x 5.33
O-ring
40.64 x 5.33
Holder
Holder
O-ring
26.57 x 3.53
O-ring
26.57 x 3.53
Ring DN50
Ring DN50
Fig 15: ISC40 sensor in screw-in
subassembly ISC40FS-SCSA
Fig 16: ISC40 sensor in weld-in
subassembly ISC40FS-SCWN
9.3 Dimensions ISC40FS
Rd 78 x 1/6
acc. DIN 405
Rd 78 x 1/6
acc. DIN 405
Rd 78 x 1/6
acc. DIN 405
2" NPT
acc. ANSI B.20.1
88 (3.46)
64 (2.52)
64 (2.52)
Ø 50
/SCWN
2" NPT
acc. ANSI B.20.1
Ø 50
/SCSA
Ø 50
/PCSA
/FCSA
Fig 17: Dimensions ISC40FS with sensor installed (dimensions in mm (inches))
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 24
03/03/16 12:58
25
9.4 Modelcode ISC40FS
Model Suffix Option Description
ISC40FS
Flow fitting subassembly
Material -F
PVDF
-P Polypropylene
-S
Stainless Steel
Process
-CS
Dairy Coupling screw-in*
connection
-CW
Dairy Coupling welded*
Thread type
-A
NPT
NPT or R
-N
No thread (for weld-in couplings)
Options
/M Material certificate 3.1. EN 10024
(for wetted metal parts only)
* Note: according to Din 11851
9.5 Spareparts ISC40FS
Parts ISC40FS subassembly
Part no. Description
Dimensions
K1541KD Holder + O-rings 2”
K1541KL Holder + O-rings 2”
K1541KA Holder + O-rings 2”
Note: Other O-ring materials are available as sparepart
Material
PVDF
Polypropylene
AISI 316 SS
O-ring
Viton
Viton
Viton
O-rings ISC40FS subassembly
Description Dimensions
Material
Quantity
O-rings -FCSA, -PCSA, -FCWN, -PCWN
K1500DF O-ring set 40.64 x 5.33; 26.57 x 3.53; 56.52 x 5.33 EPDM
5 sets
K1500DE O-ring set 40.64 x 5.33; 26.57 x 3.53; 56.52 x 5.33 Viton
5 sets
K1500DG O-ring set 40.64 x 5.33; 26.57 x 3.53; 56.52 x 5.33 Silicon
5 sets
K1500CD O-ring
40.64 x 5.33
Kalrez
1
K1500CH O-ring
26.57 x 3.53
Kalrez
1
O-rings –SCSA, -SCWN
K1500DB O-ring set 40.64 x 5.33; 26.57 x 3.53; Ring DN50 EPDM
5 sets
K1500DA O-ring set 40.64 x 5.33; 26.57 x 3.53; Ring DN50 Viton
5 sets
K1500DC O-ring set 40.64 x 5.33; 26.57 x 3.53; Ring DN50 Silicon
5 sets
K1500CD O-ring
40.64 x 5.33
Kalrez
1
K1500CH O-ring
26.57 x 3.53
Kalrez
1
K1500DD O-ring
53.34 x 5.33
Kalrez
1
Note: O-ring 53.34 x 5.33 (K1500DD) is used as a Kalrez replacement for the Ring DN50.
Part no.
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 25
03/03/16 12:58
26
10 ISC40FF – FLOW FITTING
10.1 General Specifications
10.1.1 Materials
Wetted parts
Model ISC40FF-S
Model ISC40FF-P
Model ISC40FF-F
Non-wetted parts
:
:
:
:
AISI 316 SS; Viton
Polypropylene; Viton
PVDF (Kynar); Viton
AISI 304 SS or AISI 316 SS
10.1.2 Operating range
Temperature
Model ISC40FF-S
Model ISC40FF-P
Model ISC40FF-F
Pressure
Model ISC40FF-S
Model ISC40FF-P
Model ISC40FF-F
10.1.3 Shipping details
Package size (LxWxH)
Package weight (max)
: Max. 150°C (302°F)
: Max. 100°C (212°F)
: Max. 130°C (266°F)
: Max. 10 bar (150 PSI) at operating temperature
: Max. 6 bar (90 PSI) at 20°C (68°F)
Max. 1 bar (15 PSI) at 100°C (212°F)
: Max. 10 bar (150 PSI) at 20°C (68°F)
Max. 1 bar (15 PSI) at 130°C (266°F)
: 250 x 150 x 100 mm (9.84 x 5.90 x 3.94 inch)
: 1.5 kg (3.3 lbs)
10.1.4 Process connections
1/2” NPT in flow fitting, connected with optional flange adapters to the actual installation.
Note: T
he ISC40 sensor type -GG and -TG are supplied with a Viton gasket. This gasket
may be used with the older series of ISC40FF fittings. It should be discarded when
the sensor is mounted in combination with the ISC40FF type -PA and -FA.
10.2 Installation of ISC40 sensor in ISC40FF
Installation of the flow fittings involves two steps (see Fig 18):
• Installation of the sensor
The sensor has to be mounted in the flow fitting. It is important that the position of the
sensor in the fitting allows easy flow through the hole of the sensor donut. Generally good
flow is assured if the flats of the sensor are oriented perpendicular to the outlet piping. It
is also important that the sample piping is oriented in such a way that the direction of the
flow is upwards to assure complete filling of the flow chamber
• Installation of the fitting on wall, railing or stanchion
For this purpose the flow fitting has an optional pipe/wall mounting kit /MS or /MP (see Fig
19 for details). This kit consists of a clamp ring with bolts and nuts which clamps around
the flow chamber. Therefore the flow chamber can be turned in the mounting assembly
allowing more flexibility in installation. The mounting plate can be mounted on a wall or
panel with 2-4 anchor bolts with a diameter of 10 mm (3/8”) max. For those installations
where pipe mounting is requested (2” nominal pipe), either horizontal or vertical, a saddle
and U-bolt have to be used.
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 26
03/03/16 12:58
27
70 (2.75)
70 (2.75)
O-Ring
40.64 x 5.33
Holder
Ø 10 (0.4) 4x
O-Ring
26.57 x 3.53
Ring DN50
pipe mount
(horizontal)
pipe mount
(vertical)
Fig 18: Installation of ISC40 sensor in
flow fitting
Fig 19: Pipe/wall mounting kit
Note: The ISC40 sensor type -TG does not fit in the model ISC40FF-S.
L2
127 (5)
186 (7.32)
d
10.3 Dimensions ISC40FF
ø 90 (3.55)
Option L1
/FF1
163-168
/FF2
152-157
/FF1
163-168
/FF2
152-157
/FF1
163-168
/FF2
152-157
(6.4-6.6)
(6.0-6.2)
(6.4-6.6)
(6.0-6.2)
(6.4-6.6)
(6.0-6.2)
L2
216-223
207-212
216-223
207-212
216-223
207-212
(8.5-8.8)
(8.1-8.35)
(8.5-8.8)
(8.1-8.35)
(8.5-8.8)
(8.1-8.35)
L1
Fig 20: Flow fitting ISC40FF-P, ISC40FF-F
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 27
03/03/16 12:58
Direction
of FLOW
39 (1.6)
58 (2.3)
d
158 (6.22)
28
Ø 54 (2.12)
Option L1
/FS1 269-280 (10.6-11.0)
/FS2 279-290 (11.0-11.4)
/FS3 269-280 (10.6-11.0)
/FS4 279-290 (11.0-11.4)
L2
65 (2.56)
85 (3.35)
60.5 (2.38)
79.2 (3.12)
120 (4.72)
L1
Fig 21: Flow fitting ISC40FF-S
10.4 Modelcode ISC40FF
Model Suffix Option Description
ISC40FF flow fitting
Material
-S
AISI 316 stainless steel
-P
Polypropylene (PP)
-F
PVDF (KYNAR®)
Process connection -A
NPT
1/2”NPT Flange adapters /FF1
PVDF, DN15 PN10
/FF2
PVDF, DN25 PN10
/FF3
PVDF, ANSI 1/2”-150lbs
/FF4
PVDF, ANSI 1”-150lbs
/FP1
PP, DN15 PN10
/FP2
PP, DN25 PN10
/FP3
PP, ANSI 1/2”-150lbs
/FP4
PP, ANSI 1”-150lbs
/FS1
AISI 316 SS, DN15 PN10
/FS2
AISI 316 SS, DN25 PN10
/FS3
AISI 316 SS, ANSI 1/2”- 150lbs
/FS4
AISI 316 SS, ANSI 1”- 150lbs
Mounting set
/MS
Wall/pipe for SS flow fitting
/MP
Wall/pipe for PP or PVDF flow fitting
Material certificate
/M 3.1. according EN 10024
(for wetted metal parts only)
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 28
03/03/16 12:58
29
10.5 Spareparts ISC40FF
Options ISC40FF Flow fitting, flange adapters
Part no. Description Process connection
Material
O-ring(s)
K1521AL /FF1
DN15 PN10
PVDF
Viton
K1521AP /FF2
DN25 PN10
PVDF
Viton
K1521AE /FF3
ANSI ½” – 150 lbs
PVDF
Viton
K1521AH /FF4
ANSI 1” – 150 lbs
PVDF
Viton
K1521AM /FP1
DN15 PN10
PP
Viton
K1521AQ /FP2
DN25 PN10
PP
Viton
K1521AF /FP3
ANSI ½” – 150 lbs
PP
Viton
K1521AJ /FP4
ANSI 1” – 150 lbs
PP
Viton
K1521AK /FS1
DN15 PN10
AISI 316 SS Viton
K1521AN /FS2
DN25 PN10
AISI 316 SS Viton
K1521AD /FS3
ANSI ½” – 150 lbs
AISI 316 SS Viton
K1521AG /FS4
ANSI 1” – 150 lbs
AISI 316 SS Viton
Note: Other O-ring materials are available as sparepart
Parts ISC40FF Flow fitting
Part no. Description
Dimensions
K1541KD Holder + O-rings 2”
K1541KL Holder + O-rings 2”
K1541KA Holder + O-rings 2”
Note: Other O-ring materials are available as sparepart
Material
O-ring
PVDF
Viton
Polypropylene Viton
AISI 316 SS Viton
O-rings ISC40FF Flow fitting
Description Dimensions
Material
Quantity
O-rings -SA
K1500DB O-ring set 40.64 x 5.33; 26.57 x 3.53; Ring DN50 EPDM
5 sets
K1500DA O-ring set 40.64 x 5.33; 26.57 x 3.53; Ring DN50 Viton
5 sets
K1500DC O-ring set 40.64 x 5.33; 26.57 x 3.53; Ring DN50 Silicon
5 sets
K1500CD O-ring
40.64 x 5.33
Kalrez
1
K1500CH O-ring
26.57 x 3.53
Kalrez
1
K1500DD O-ring
53.34 x 5.33
Kalrez
1
O-rings -PA, -FA
K1500DF O-ring set 40.64 x 5.33; 26.57 x 3.53; 56.52 x 5.33 EPDM
5 sets
K1500DE O-ring set 40.64 x 5.33; 26.57 x 3.53; 56.52 x 5.33 Viton
5 sets
K1500DG O-ring set 40.64 x 5.33; 26.57 x 3.53; 56.52 x 5.33 Silicon
5 sets
K1500CD O-ring
40.64 x 5.33
Kalrez
1
K1500CH O-ring
26.57 x 3.53
Kalrez
1
Note: O-ring 53.34 x 5.33 (K1500DD) is used as a Kalrez replacement for the Ring DN50.
Part no.
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 29
03/03/16 12:58
30
11 ISC40FD – IMMERSION FITTING
11.1 General Specifications
11.1.1 Materials
Wetted parts without options
Model ISC40FD-S
: AISI 316 SS, Viton
Model ISC40FD-V
: C-PVC, Viton
Flange ISC40FD-S-**-SF*
: AISI 316 SS
Non-wetted parts
Tube pigtail
: plastic
11.1.2 Operating range
Temperature
Model ISC40FD-S
Model ISC40FD-V
Pressure
Model ISC40FD-S
Model ISC40FD-V
: Max. 150°C (302°F)
: Max. 80°C (176°F)
: Max. 10 bar (150 PSI) at operating temperature
: Max. 2 bar (30 PSI) at 20°C (68°F)
Max. 1 bar (15 PSI) at 80°C (176°F)
11.1.3 Process connections
Flange AISI 316 SS 2” or AISI 316 SS DN50, for the type –V an adjustable optional flange
is available with ANSI 2” 150 lbs and DN50 PN10 hole pattern.
Note: T
he ISC40 sensor type -GG and -TG are supplied with a Viton gasket. This gasket
may be used with the older series of ISC40FD fittings. It should be discarded when
the sensor is mounted in combination with the ISC40FD type -S and -V.
11.2 Installation of ISC40 sensor in ISC40FD
Installation of the immersion fittings involves two steps:
• Installation of the sensor
The sensor has to be mounted in the immersion fitting (see Fig 22).
• Installation in the application
o On a railing or stanchion
For this purpose the immersion fitting has an optional rail mounting kit /MS1. This kit
consists of a piece of guide pipe and two pipe clamps. The guide pipe is mounted
horizontally to the stanchion using one pipe clamp. The second pipe clamp connects
the guide pipe to the immersion fitting. The distance between the pipe clamps
determines the distance between the stanchion and immersion fitting (see Fig 23).
oW
ith flange
For this purpose a stainless steel immersion fitting can be ordered as type SFD (DN50) or type –SFA (2”). The PVC immersion fitting has an optional depth
adjustable mounting flange /FA (with DN50 and ANSI 2” 150 lbs hole pattern) which
is illustrated in Fig 24. You must assure that the mating flange has the same hole
pattern as the supplied flange.
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 30
03/03/16 12:58
31
ISC40FD-S
ISC40FD-V
Guide pipe
Ø 48,6 x 600
(Ø1.9 x 24")
Min: 110 (4.35")
Max: 515 (20")
O-Ring
44.17 x 1.78
O-Ring
39.35 x 2.62
2-inch Stanchion
(Vertical)
O-Ring
26.57 x 3.53
Fig 22: Installation of ISC40 sensor
in immersion fitting.
Fig 23: Installation on stanchion
with /MS1
Fig 24: Option /FA
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 31
03/03/16 12:58
32
Alternative ways of mounting the immersion fitting are:
• Guide pipe
To facilitate this type of mounting the immersion fitting has a larger diameter at the top.
The user supplied guide pipe (internal diameter > 52 mm) is fitted to the wall or mounting
rail. The immersion fitting slides into this guide pipe and therefore the sensor can be easily
removed for inspection.
• Platform mounting
Sometimes there is a walking platform that can be used to mount the immersion fitting.
The top of the immersion fitting has a larger diameter than the shaft. The only thing to do
is to drill a precise hole in the platform to slide the fitting through.
11.3 Dimensions ISC40FD
Dimensions in mm (inches)
42 (1.65)
Transmitter side
ø60 (2.5)
Locknut
ø50 (2)
Sealing
L*
ø51 (2)
Sealing
L* see modelcode for available insertion
lengths
Fig 25: Immersion fitting ISC40FD
Fitting side
Fig 26: P
rotection hose,
option /PH5, /PH10
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 32
03/03/16 12:58
33
11.4 Modelcode ISC40FD
Model code Suffix code Option code Description
ISC40FD Immersion fitting
Material
-S
AISI 316 Stainless steel
-V PVC-C
Insertion length Between 05 to 20
Example: 05 = 0.5 m / 20 = 2.0 m
Flange
-NFL
No flange
-SFA
AISI316 SS 2”
-SFD
AISI316 SS DN50
Options
/MS1
Pipe mounting set (Carbon steel)
/FA
Adjustable flange with DIN DN50-PN10 and
ANSI 2” 150 lbs hole pattern (only for PVC)
/PH5
Protection hose for 5 m cable
/PH10
Protection hose for 10 m cable
Material certificate
/M 3.1. according EN 10204
(for wetted metal parts only)
Note: option /FA only for ISC40FD-V; option /M only for ISC4FD-S
11.5 Spareparts ISC40FD
Options ISC40FD Flow fitting
Description Process connection
/MS1
/FA
DN50 PN10; ANSI 2” lbs
/PH5
/PH10
O-rings ISC40FD Immersion fitting
Part no. Description Dimensions
O-rings -S
K1500CE O-ring set 39.35 x 2.62; 26.57 x 3.53
K1500CF O-ring set 39.35 x 2.62; 26.57 x 3.53
K1500CG O-ring set 39.35 x 2.62; 26.57 x 3.53
K1500CH O-ring
26.57 x 3.53
O-rings -V
K1500CX O-ring set 39.35 x 2.62; 26.57 x 3.53; 44.17 x 1.78
K1500CW O-ring set 39.35 x 2.62; 26.57 x 3.53; 44.17 x 1.78
K1500CY O-ring set 39.35 x 2.62; 26.57 x 3.53; 44.17 x 1.78
Part no.
K1541ZY
K1520EV
K1500CJ
K1500CK
Material
Carbon steel
PP
PVC, PA6.6
PVC, PA 6.6
Material
Quantity
EPDM
Viton
Silicon
Kalrez
5 sets
5 sets
5 sets
1
EPDM
Viton
Silicon
5 sets
5 sets
5 sets
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 33
03/03/16 12:58
34
12 Chemical Compatibility Chart
Basic
salt
Neutral
salt
Oxidizing
agent
Organic
solvent
O = can be used, X = shortens useful life, - = cannot be used
Acid
salt
Alkali
Organic acid
Inorganic acid
Sulfuric acid
10
50
95
fuming
Hydrochloric acid
10
sat.
Nitric acid
25
50
95
fuming
Phosphoric acid
25
50
95
Hydrofluoric acid
40
75
Acetic acid
10
glacial
Formic acid
80
Citric acid
50
Calcium hydroxide
sat.
Potassium hydroxide
50
Sodium hydroxide
40
Ammonia in water
30
Ammonium chloride
sat.
Zinc chloride
50
Iron (III) chloride
50
Sodium sulfite
sat.
Sodium carbonate
sat.
Potassium chloride
sat.
Sodium sulfate
sat.
Calcium chloride
sat.
Sodium chloride
sat.
Sodium nitrate
50
Aluminium chloride
sat.
Hydrogen peroxide
30
Sodium hypochloride
50
Potassium dichromate sat.
Chlorinated lime
Ethanol
80
Cyclohexane
Toluene
Trichloroethane
Water
O O O X X X O O O O O O O O
O O O X X X O O O O O X O O
OX - XXX OOO - - - X -
- - - - - - OOO - - - - -
OOO - - - OOO OOX OO
O O O - - -
OOX OO
O O X X X X O O X O O O O O
OOX XXX - - - XXX X -
OX - OOO - - - - - - - -
- - - OOO - - - - - - - -
OOO - - - OOO OOO OO
O O O X X X O O O O O O O O
OOO OOO XX - OOO OO
OOO - - - OOO - - - OO
OOO - - - OOO - - - OO
OOO OOX - - - OOO OO
OX - OOX - - - OOX OX
OOO XXX - - - XXX OO
OOO OOO OOO OOO OO
O O O O O O O O O O O O O O
O O X O O O O O O O O O O O
O O X O O O X X X O O O O O
OOO OOO XXX OOO OO
O O O X X X O O O O O O O O
O O O X X X O O O O O O O O
OOO - - - OOO OOO OO
OOO OOO - - - OOO OO
O O O O O O O O O O O O O O
O O O X X X O O O O O O O O
OOO OOO OOO OOO OO
O O O X X X O O O O O O O O
O O O X X X O O O O O O O O
O O O X X X O O O O O O O O
OOO - - - OOO OOO OO
O O O O O O O O O O O O O O
OOO XXX OOX OOO XX
O O O O O O O O O O O O O O
O X - X X X
XXX - -
OOX OOO X - - OOO OO
OOX OOO OOO OOO - -
OOO OOO - - - OOO X -
XXX OOX XXX OOO - -
OOO OOO OOO OOO OO
PFA
20
60
100
20
60
Material
PP
PVC
20
60
20
60
100
20
60
100
20
60
100
Temp.
%
ºC
Conc.
20
60
100
PVDF S.S. 316 VITON PEEK
(Kynar)
OX
OO
XX
- -
OX
OO
OX
OX
- -
- -
OX
OO
OO
OX
XX
OX
XX
O-
OO
OO
OO
OX
OX
OO
OO
OO
OO
OO
OO
OO
OO
OO
OO
OO
OO
XX
OO
OO
OX
OO
- -
- -
OO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOX
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
OOO
Note: Information in this list is based on our general experience and literature data and
given in good faith. However Yokogawa is unable to accept responsobility for claims
related to this information.
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 34
03/03/16 12:58
35
IM 12D08J02-E-E
IM 12D08J02-E-E_ed18.indd 35
03/03/16 12:58
YOKOGAWA ELECTRIC CORPORATION
World Headquarters
9-32, Nakacho 2-chome, Musashino-shi
Tokyo 180-8750
Japan
www.yokogawa.com
YOKOGAWA ELECTRIC ASIA Pte. LTD.
5 Bedok South Road
Singapore 469270
Singapore
www.yokogawa.com/sg
YOKOGAWA CORPORATION OF AMERICA
2 Dart Road
Newnan GA 30265
USA
www.yokogawa.com/us
YOKOGAWA CHINA CO. LTD.
3F Tower D Cartelo Crocodile Building
No.568 West Tianshan Road Changing District
Shanghai, China
www.yokogawa.com/cn
YOKOGAWA EUROPE BV
Euroweg 2
3825 HD AMERSFOORT
The Netherlands
www.yokogawa.com/eu
YOKOGAWA MIDDLE EAST B.S.C.(c)
P.O. Box 10070, Manama
Building 577, Road 2516, Busaiteen 225
Muharraq, Bahrain
www.yokogawa.com/bh
IM 12X0X0-E-E
Subject to change without notice
Copyright ©
IM 12D08J02-E-E
Subject to change without notice
Copyright ©
IM 12D08J02-E-E_ed18.indd 36
Yokogawa has an extensive sales and
distribution network.
Please refer to the European website
(www.yokogawa.com/eu) to contact your
nearest representative.
Printed in The Netherlands, 00-000 (A) I
Printed in The Netherlands, 18-1603 (A) I
03/03/16 12:58