Endress + Hauser Modular Analytical Panel System with Sample preparation for Steam and Power Applications
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Special Documentation
Modular Analytical Panel System with Sample preparation
for Steam and Power Applications
Catalogue and Operating Instructions
SD_Modular_Panels_Power_V2.0.doc/04_12
Modular Analytical Panel System with Sample Conditioner
Table of Contents:
1
1.1
1.2
1.3
1.4
1.5
2
SAFETY INSTRUCTIONS..................................................................................................................4
DESIGNATED USE ..............................................................................................................................4
INSTALLATION, COMMISSIONING AND OPERATION .........................................................................4
OPERATIONAL SAFETY.......................................................................................................................5
RETURN .............................................................................................................................................5
SAFETY ICONS ....................................................................................................................................5
OVERVIEW OF AVAILABLE PANELS ............................................................................................6
2.1 GENERAL INFORMATION ...................................................................................................................6
2.2 SAMPLE PREPARATION PANELS SL-100-L/H – C-A070813-71/72 SL-200-L/H – CA110121-83/84 .......................................................................................................................................7
2.2.1 SINGLE LINE SAMPLE PANEL SL100L/H – C-A070813-71/72..........................................................8
2.2.2 SINGLE LINE SAMPLE PANEL SL200L/H – C-A110121-83/84..........................................................9
2.3 SAMPLE PREPARATION PANELS PE 3A C-A110329-82 PE 2A C-A100303-90 ...................10
2.3.1 SAMPLE PREPARATION PANEL PE 3A (WATER) C-A110329-82........................................................13
2.3.2 SAMPLE PREPARATION PANEL PE 2A (STEAM) C-A100303-90 ........................................................14
2.4 SAMPLE PREPARATION PANELS KEP02 C-A120327-80 KEP05-4 C-A120327-81 KEP058 C-A120327-82 ....................................................................................................................................15
2.4.1 SAMPLE PREPARATION PANEL KEP02 (WATER) C-A120327-80 ...............................................16
2.4.2 SAMPLE PREPARATION PANEL KEP05-4 (STEAM) C-A120327-81.............................................17
2.4.3 SAMPLE PREPARATION PANEL KEP05-8 (STEAM) C-A120327-82.............................................18
2.5 CONDUCTIVITY / DISSOLVED OXYGEN / PH PANEL C-A100120-70 / -71 ................................19
2.6 CONDUCTIVITY PANEL
C-A100617-81 ....................................................................................20
2.7 CONDUCTIVITY + DISSOLVED OXYGEN OR CONDUCTIVITY + PH C-A100208-50 .......................21
2.8 DISSOLVED OXYGEN + PH C-A101222-80 ...............................................................................22
2.9 CATIONIC AND DIFFERENTIAL CONDUCTIVITY PANEL C-A120327-83........................................23
2.10 CATIONIC AND DIFFERENTIAL CONDUCTIVITY PANEL C-A110324-80 / 81 .............................24
2.11 CA71 ANALYZER PANEL
C-A100303-91 ................................................................................25
2.12 SODIUM (NA) ANALYZER PANEL C-A100303-92
PRELIMINARY DESIGN!...........................26
2.13 ANALYZER FOR DEGASSED CATIONIC CONDUCTIVITY C-A110318-80.......................................27
2.14 ANALYZER FOR DEGASSED CATIONIC CONDUCTIVITY C-A120327-84.......................................28
3
3.1
3.2
3.3
4
4.1
4.2
5
5.1
ACCESSORIES FOR MOUNTING PLATES..............................................................................29
ALUMINIUM PROFILE FRAME C-A100202-50...........................................................................29
WEATHER PROTECTION ROOF C-A100202-51.........................................................................30
DRAIN GUTTER
C-A100202-53...............................................................................................31
IDENTIFICATION ...........................................................................................................................32
NAME PLATE ...................................................................................................................................32
SCOPE OF DELIVERY .........................................................................................................................32
INSTALLATION...............................................................................................................................33
INCOMING ACCEPTANCE, TRANSPORT, STORAGE ..........................................................................33
Endress+Hauser
2
Modular Analytical Panel System with Sample Conditioner
5.2
6
INSTALLATION REQUIREMENTS .......................................................................................................33
CONNECTIONS...............................................................................................................................34
6.1
6.2
6.3
ELECTRICAL CONNECTION .............................................................................................................34
MEDIUM CONNECTION ...................................................................................................................34
CONNECTION BETWEEN TWO OR MORE PANELS .............................................................................34
7
OPERATION.....................................................................................................................................35
8
MAINTENANCE ..............................................................................................................................37
9
GENERAL ACCESSORIES...............................................................................................................38
10
TROUBLE SHOOTING..................................................................................................................39
11
TECHNICAL DATA .......................................................................................................................40
11.1
PROCESS CONDITIONS...................................................................................................................40
Endress+Hauser
3
Modular Analytical Panel System with Sample Conditioner
1
Safety Instructions
1.1 Designated Use
The Modular Analytical Mounting Plate System is a sampling system for measuring pH, ORP,
conductivity, cationic conductivity or dissolved oxygen in water and steam circuits of power
plants in NON-EX area.
Depending on the used Sample Preparation Panel up to three panels can be easily connected to a
full analytical system. See chapter 2.2 and 2.3 for further details.
The mounting panels shall be wall mounted in dry rooms protected from frosty ambient conditions. They can also be installed in monitoring containers.
Main area of application:
Quality measurements in water and steam circuits of power plants, heating plants and incineration plants.
Features and benefits:
•
•
•
•
•
•
•
•
•
•
compact, vertical arrangement and easy connection of plates
reliable operation as special sensor properties are taken into account in the layout.
choice of different sample preparation panels for different demands
max. pressure at sample inlet: up to 245 bar depending on the sample panel
max. temperature at sample inlet: up to 550°C depending on the sample panel
optional thermal shut off valve in sample line after cooler to protect operating personal and analyzers
optional flow meter with switching contact or flow indicator for cooling water circuit
adjustable flow rate for every measurement parameter
optional flow meter with switching contact at every sample line
integration of Third Party components in the system on request
1.2 Installation, Commissioning and Operation
Please note the following items:
• Installation, commissioning, operation and maintenance of the measuring system
must only be carried out by trained technical personnel.
• The technical personnel must be authorized for the specified activities by the system operator.
• Electrical connection must only be carried out by a certified electrician.
• Technical personnel must have read and understood these Operating Instructions
and must adhere to them.
• Before commissioning the entire measuring point, check all the connections for
correctness. Ensure that electrical cables and hose connections are not damaged.
• Before commissioning check all fittings and retighten them if necessary to avoid
leakages.
• Do not operate damaged products and secure them against unintentional commissioning. Mark the damaged product as being defective.
Endress+Hauser
4
Modular Analytical Panel System with Sample Conditioner
•
•
•
Measuring point faults may only be rectified by authorized and specially trained
personnel.
If faults can not be rectified, the products must be taken out of service and secured against unintentional commissioning.
Repairs are not described in these Operating Instructions and may only be carried
out at the manufacturer or by the service organisation.
1.3 Operational Safety
The Analytical Panel has been designed and tested according to the state of the art and left the
factory in perfect functioning order.
Relevant regulations and European standards have been met.
As the user, you are responsible for complying with the following safety conditions:
•
•
Installation instructions
Local prevailing standards and regulations.
EMC
The instruments mounted on the panels have been tested for electromagnetic compatibility in
industrial use according to applicable European standards. Protection against interference as
specified above is valid only for an instrument connected according to the instructions in these
Standard Operating Instructions.
1.4 Return
If a mounting plate has to be repaired, please return it cleaned to the sales center responsible.
Please use the original packaging, if possible.
1.5
#
"
!
Safety Icons
Warning!
This symbol alerts you to hazards. They can cause serious damage to the instrument or to
persons if ignored.
Caution!
This symbol alerts you to possible faults which could arise from incorrect operation. They
could cause damage to the instrument if ignored.
Note!
This symbol indicates important items of information.
Endress+Hauser
5
Modular Analytical Panel System with Sample Conditioner
2
Overview of Available Panels
2.1
General Information
Tubing:
Stainless Steel tubing (316 or 316L): 25 mm or ½ inch for cooling
water lines, 12 mm or ¼ inch sample in- and outlet lines, 6 mm sample lines to flow vessels, Swagelok fittings (316) and clamping ring
connections
Mounting plates:
Stainless Steel (304) or partially powder coated aluminum
Media:
Boiler feed water, condensate, heating water, steam etc.
Dimensions:
Sample preparation plate Sentry SL100/200:
H x W x D: 1000 x 305mm
Sample preparation plate Thiedig PE02/PE03:
H x W x D: 1000 x 450mm
Analysis plates: H x W x D: 1000 x 300mm
Sample preparation plate Woesthoff KEP02/KEP05:
H x W x D: 1000 x 500mm
Analysis plates: H x W x D: 1000 x 400mm
Features:
!
Endress+Hauser
Sensor and transmitter – selectable for best performance – must be
ordered separately with a suitable cable
Depending on the type of Sample Preparation Panel up to three panels can easily be connected. Flow meters or flow guards with
switching contact can be ordered as an option. This allows to
adjust an appropriate flow on each panel.
Note!
All standard components like sensors, assemblies (if mentioned), cables, transmitters etc. are not included in the scope of delivery of the mounting plate. They
have to be ordered separately!
6
Modular Analytical Panel System with Sample Conditioner
2.2
Sample Preparation Panels SL-100-L/H – C-A070813-71/72
SL-200-L/H – C-A110121-83/84
!
Note!
The panel containing the cooler and the pressure reducing device must be used
always as the first panel in a row.
These panels are intended to be used for different pressure and temperature sample
conditions, primary for water or condensate. Depending on the type, one (with SL100)
respectively up to three (with SL200) analytical panels can be attached.
Figure 1.: Example of a sample conditioning panel with different options (exemplary, may vary
with the order), brand Sentry
Endress+Hauser
7
Modular Analytical Panel System with Sample Conditioner
2.2.1 Single Line Sample Panel SL100L/H – C-A070813-71/72
!
Note!
The panel containing the cooler and the pressure reducing device must
be always used as the first panel in a row. The designated use is grab
sample only, it can be used with one further attached analytical panel
with low (L-version) and high (H-version) pressure and temperature
sample inlet conditions after check up in particular case.
optional
optional
optional
optional
Figure 2.: P&I scheme of the sample conditioning panel SL100L/H with different options (may
vary with the order).
Ordering Information
Process conditions
Sample Inlet
51515151 + 71181119
SL100L
Pmax. = 34 bar; Tmax. = 232 °C
51515151 + 71180834
SL100H
Pmax. = 35 - 209 bar; Tmax. = 538 °C
For further details and optional components please refer to the Single Line Sample Panel Manual
SPD 1.1.4 which is available in the TSP modification database.
The process data of the cooling water as pressure, Tin, Tout/max and Flowrate must be
provided. This defines the operational range and the type of the cooler.
Endress+Hauser
8
Modular Analytical Panel System with Sample Conditioner
2.2.2 Single Line Sample Panel SL200L/H – C-A110121-83/84
!
Note!
The panel containing the cooler and the pressure reducing device must
be always used as the first panel in a row. It can be used with a max. of
three further attached analytical panel with low (L-version) and high (Hversion) pressure and temperature sample inlet conditions.
optional
optional
optional
optional
optional
optional
optional
optional
Figure 3.: P&I scheme of the sample conditioning panel SL200L/H with different options (may
vary with the order).
Ordering Information
51515151 + 71178791
51515151 + 71180881
SL200L
SL200H
Process conditions
Sample Inlet
Pmax. = 34 bar; Tmax. = 232 °C Pmax. = 35 - 209 bar; Tmax. = 538 °C
For further details and optional components please refer to the Single Line Sample Panel Manual
SPD 1.1.4 which is available in the TSP modification database.
The process data of the cooling water as pressure, Tin, Tout/max and Flowrate must be
provided. This defines the operational range and the type of the cooler.
Endress+Hauser
9
Modular Analytical Panel System with Sample Conditioner
2.3
Sample preparation panels PE 3A
PE 2A
!
C-A110329-82
C-A100303-90
Note!
The panel containing the cooler and the pressure reducing device must be always
used as the first panel in a row.
cooling water sight glass
(option)
Thermometer
protection insulation
inlet blow down valve
(option)
temperature shut off
valve (option)
back pressure regulator
(option)
flow guard
(option)
back plate coated
aluminium
connection to analysis
panel (Swagelok compression fitting for
12mm tube)
Figure 4.: Example of a high temperature sample preparation panel with steam cooler, brand
Dr. Thiedig, with optional accessories
Endress+Hauser
10
Modular Analytical Panel System with Sample Conditioner
Features and benefits:
•
•
•
•
•
high pressure inlet valve, welded at sample inlet and with protection cover
needle valve for pressure reduction to ensure constant sample flow to analysis
a backpressure regulator is available as option
cooler material 1.4571 or 1.4563 for high temperature application
aluminium back plate with powder coating
Available documentation on request (with additional costs):
•
•
•
•
•
•
order related drawings
certificate of compliance with the order (EN10204 - 2.1)
pressure test and functional test for the high-pressure valves (EN10204 – 3.1)
material certificate for the valve body (EN10204 – 3.1)
pressure test for the cooler (EN10204 – 3.1)
material certificate for the coil (EN10204 – 3.1)
Advantages:
High-Pressure Sample Cooler
•
•
•
•
•
•
•
•
higher cooling capacity, reduced pressure drop in the sample line and cooling water consumption
the capacity of the coolers is designed to 60 l/h in accordance with VGB 450
coolers are fabricated free of brass and are 100% austenitic steel
counter flow principle for cooling water and sample
cooling water flow of 1,6m³/h (steam cooler) and 1,1m³/h (water cooler) leads to sample
outlet temperature of 5-10K above the cooling water inlet temperature
cooler can be maintained and cleaned in a short time without complete dismounting in
place
coolers can be opened without de-installation
cooling coils made of thermal and chemical resistant Austenite 1.4563 (Alloy 28) for
temperatures > 400°C
High-Pressure Valves
•
•
•
•
•
•
•
•
robust needle valves for cold pressure reduction up to 320 bar with 200°C (PN400)
robust inlet valves with differential thread for handling
minimized wear and secure shut-off by non-turning spindle
thread of the spindle is not in contact with the sample media
identical internal diameters of inlet valve and inlet sample line, this reduces pressure gap
and fouling
stuffing boxes can be adjusted without opening the valves
forged valve bodies and corrosion resistant spindles lead to extended life-time
valve bodies made of heat resistant Austenite 1.4988 for temperature > 400°C
Endress+Hauser
11
Modular Analytical Panel System with Sample Conditioner
System and material of panels PE02 / PE03
•
•
•
•
•
•
•
•
•
•
•
•
•
•
equipment for inlet temperatures of up to 630°C
all high pressure components are tested separately
only designed for hot blow down before coolers to achieve the required capacities and acceleration
standards for bending ranges according design temperature and pressure are recognized
Made in Germany
high pressure components seal metal to metal for stable sealing
low pressure gap in the sampling line for optimal supply of the analyzers
system design according to VGB 450L and ASTM designation
design according ergonomic principles for improved safety of operation personal
touch guard for lines and valve bodies upstream cooling
welded inlet lines, valves and coolers
welding testing according AD 2000 and EN 13480 / 13445 possible
wear parts of the valves exchangeable without demounting
Austenite 1.4571 (316 Ti) for the high pressure tubing and all high pressure components
in contact to the sample
Endress+Hauser
12
Modular Analytical Panel System with Sample Conditioner
2.3.1 Sample preparation panel PE 3A (water)
cooling water inlet valve
C-A110329-82
thermometer
pressure reducing valve
– > needle valve for
constant flow
backplate aluminum;
stainless steel as option
Figure 5.: sample preparation panel with water cooler PE03 (without additional options)
Ordering Information
Optional
51515151 + 71180908
cooling water sight glass, flow guard, inlet blow down valve, back pressure regulator, temperature shutoff valve @ 55°C, backplate stainless
steel
Process conditions
Sample inlet
Pmax. = 245 bar; Tmax. = 250°C
cooling water inlet
Pmax. = 16 bar; Tmax. = 100 °C
All high pressure tubes are welded, low pressure tubes are connected with Swagelok compression
fittings. Welding certificates according to the European Pressure Equipment Directive 97/23/EG
and DIN ISO 13445 plus inspection certificates EN10204 3.1 can be ordered additionally.
The process data of the cooling water as pressure, Tin, Tout/max and Flowrate must be
provided. This defines the operational range.
accessories please see chapter 3
Endress+Hauser
13
Modular Analytical Panel System with Sample Conditioner
2.3.2 Sample preparation panel PE 2A (steam)
C-A100303-90
option:
cooling water sight glass
option:
inlet blow down
valve
backplate coated
aluminium;
stainless steel as
option
option:
temperature shutoff
valve (55°C)
option:
backpressure regulator
option:
flow guard with limit
switch
Figure 6.: sample preparation panel with steam cooler PE02 (picture shows additional options)
Ordering Information
Optional
51515151 + 71180341
cooling water sight glass, flow guard, inlet blow down, back pressure
regulator, temperature shutoff valve @ 55°C, backplate stainless steel,
high temperature version
Process conditions
Sample inlet
Pmax. = 245 bar; Tmax. = 400°C /
Pmax. = 245 bar; Tmax. = 550°C (high temperature version)
cooling water inlet
Pmax. = 16 bar; Tmax. = 100 °C
All high pressure tubes are welded, low pressure tubes are connected with Swagelok compression
fittings. Welding certificates according to the European Pressure Equipment Directive 97/23/EG
and DIN ISO 13445 plus inspection certificates EN10204 3.1 can be ordered additionally.
The process data of the cooling water as pressure, Tin, Tout/max and Flowrate must be
provided. This defines the operational range.
accessories please see chapter 3
Endress+Hauser
14
Modular Analytical Panel System with Sample Conditioner
2.4
Sample preparation panels KEP02
KEP05-4
KEP05-8
!
C-A120327-80
C-A120327-81
C-A120327-82
Note!
The panel containing the cooler and the pressure reducing device must be always
used as the first panel in a row.
sample inlet, Option:
2nd and 3rd inlet valve
Options: cooling
water sight glass;
sample filter; blow
down valve
Options: flow meter;
mechanical temperature shutoff valve;
pressure gauge
back panel
TRESPA; Option:
stainless steel
Option: pressure
relief valve in cooling water inlet
Option: backpressure regulator with
manual grab sample
Figure 7.: Example of a high temperature sample preparation panel with steam cooler, brand
Woesthoff, with optional equipment
Endress+Hauser
15
Modular Analytical Panel System with Sample Conditioner
2.4.1 Sample preparation panel KEP02 (water)
C-A120327-80
Figure 8.: sample preparation panel with water cooler G02S (picture shows additional options)
Ordering Information
Optional
51515151 + 71180341
e.g. backplate stainless steel, cooling water sight glass, flow guard, inlet
blow down, back pressure regulator, temperature shutoff valve @
50°C, second sample inlet valve (please see TSP data base for full options list)
Process conditions
Sample inlet
Pmax. = 65 bar; Tmax. = 200°C
cooling water inlet
Pmax. = 16 bar; Tmax. = 100 °C
All tubes are connected with Swagelok compression fittings. Inspection certificates EN10204 3.1
can be ordered additionally.
The process data of the cooling water as pressure, Tin, Tout/max and Flowrate must be
provided. This defines the operational range.
accessories please see chapter 3
Endress+Hauser
16
Modular Analytical Panel System with Sample Conditioner
2.4.2 Sample preparation panel KEP05-4 (steam)
C-A120327-81
Figure 9.: sample preparation panel with water cooler G05S-4 (picture shows additional options)
Ordering Information
Optional
51515151 + 71184184
e.g. backplate stainless steel, cooling water sight glass, flow guard, inlet
blow down, back pressure regulator, temperature shutoff valve @
50°C, second sample inlet valve (please see TSP data base for full options list)
Process conditions
Sample inlet
Pmax. = 160 bar; Tmax. = 400°C /
Pmax. = 320 bar; Tmax. = 400°C (high pressure version)
cooling water inlet
Pmax. = 16 bar; Tmax. = 100 °C
All tubes are connected with Swagelok compression fittings. Inspection certificates EN10204 3.1
can be ordered additionally.
The process data of the cooling water as pressure, Tin, Tout/max and Flowrate must be
provided. This defines the operational range.
accessories please see chapter 3
Endress+Hauser
17
Modular Analytical Panel System with Sample Conditioner
2.4.3 Sample preparation panel KEP05-8 (steam)
C-A120327-82
Figure 10.: sample preparation panel with water cooler G05S-8 (picture shows additional options)
Ordering Information
Optional
51515151 + 71184220
e.g. backplate stainless steel, cooling water sight glass, flow guard, inlet
blow down, back pressure regulator, temperature shutoff valve @
50°C, second sample inlet valve (please see TSP data base for full options list)
Process conditions
Sample inlet
Pmax. = 250 bar; Tmax. = 550°C /
Pmax. = 250 bar; Tmax. = 650°C (high temperature version)
cooling water inlet
Pmax. = 16 bar; Tmax. = 100 °C
All tubes are connected with Swagelok compression fittings. Inspection certificates EN10204 3.1
can be ordered additionally.
The process data of the cooling water as pressure, Tin, Tout/max and Flowrate must be
provided. This defines the operational range.
accessories please see chapter 3
Endress+Hauser
18
Modular Analytical Panel System with Sample Conditioner
2.5
Conductivity / Dissolved Oxygen / pH panel
C-A100120-70 / -71
Pmax=4bar
Tmax=80°C
Figure 11.: Conductivity/DO panel.
Ordering Information
Sensor
Cable
Assembly
Transmitter
Optional
VA-Flow meter for sample
See chapter 2.1 for further details.
51515151 + 71180887
Conductivity
CLS15D-#1A1
51515151 + 71180819
Dissolved Oxygen or pH
COS21D-C11## or
CPS11D-7AS21 or
CPS41D-7AB2B1 with
CPY7-B0 (not included)
CYK10-A881 (1,5m)
CYK10-A881 (1,5m)
Included (SAP 71042405)
Included (SAP 71042404)
CM442-###########
0,1 – 0,6 L/min, switching contact (Non-Ex, 200 V, 1 A, 20
VA, IP65)
accessories please see chapter 3
Endress+Hauser
19
Modular Analytical Panel System with Sample Conditioner
2.6
Conductivity panel
C-A100617-81
Pmax=4bar
Tmax=80°C
Figure 12.: Conductivity panel.
Ordering Information
Sensor
Cable
Assembly
Transmitter
Optional
VA-Flow meter for sample
See chapter 2.1 for further details.
51515151 + 71179559
Conductivity
CLS21D-C1E1
CYK10-A881 (1,5m)
Included (SAP 50012519)
CM442-###########
0,1 – 0,6 L/min, switching contact (Non-Ex, 200 V, 1 A, 20
VA, IP65)
accessories please see chapter 3
Endress+Hauser
20
Modular Analytical Panel System with Sample Conditioner
2.7
Conductivity + Dissolved Oxygen or Conductivity + pH
C-A100208-50
Pmax=4bar
Tmax=80°C
Figure 13.: Conductivity / DO or conductivity / pH panel.
Ordering Information
Sensor
Cable
Assembly
Transmitter
Optional
VA-Flow meter for sample
See chapter 2.1 for further details.
51515151 + 71181048
Conductivity
Dissolved Oxygen or pH
CLS15D-#1A1
COS21D-C11## or
CPS11D-7AS21 or
CPS41D-7AB2B1 with
CPY7-B0 (not included)
CYK10-A881 (1,5m)
CYK10-A881 (1,5m)
Included (SAP 71042405)
Included (SAP 71042404)
CM442-##M2#######
0,1 – 0,6 L/min, switching contact (Non-Ex, 200 V,
20 VA, IP65)
1 A,
accessories please see chapter 3
Endress+Hauser
21
Modular Analytical Panel System with Sample Conditioner
2.8
Dissolved Oxygen + pH
C-A101222-80
Pmax=4bar
Tmax=80°C
Figure 14.: DO / pH panel
Ordering Information
Sensor
Cable
Assembly
Transmitter
Optional
VA-Flow meter for sample
See chapter 2.1 for further details.
51515151 + 71178418
Dissolved Oxygen
pH
COS21D-C11##
CPS11D-7AS21 or
CPS41D-7AB2B1 with
CPY7-B0 (not included)
CYK10-A881 (1,5m)
CYK10-A881 (1,5m)
Included (SAP 71042404)
Included (SAP 71042404)
CM442-##M2#######
0,1 – 0,6 L/min, switching contact (Non-Ex, 200 V,
20 VA, IP65)
1 A,
accessories please see chapter 3
Endress+Hauser
22
Modular Analytical Panel System with Sample Conditioner
2.9
Cationic and Differential Conductivity panel
C-A120327-83
CM442
(not included)
backplate Trespa;
stainless steel as
option
CLS15D
(not included)
flow indicator as option
cation filter with assembly for
2x CLS15D in price included
Pmax=4bar
Tmax=55°C
free outlet
Figure 15.: Cationic and Differential Conductivity panel.
Ordering Information
51515151 + 71184222
Sensor
2 x CLS15D-#1A1 for differential conductivity
Cable
Assembly
2 x CYK10-A881 (1,5m)
Included in cation filter (0.6 litre volume); resin included;
flow indicator in cation filter included
CM442-##M2#######
backplate stainless steel, larger filter housing for 1.2L resin,
flow indicator
Transmitter
Option
accessories please see chapter 3
Endress+Hauser
23
Modular Analytical Panel System with Sample Conditioner
2.10 Cationic and Differential Conductivity panel
C-A110324-80 / 81
CM442
(not included)
backplate coated
aluminium;
stainless steel as
option
CLS15D
(not included)
cation filter with flow indicator
and assembly for 2x CLS15D in
price included
Pmax=3,2bar
Tmax=60°C
free outlet
Figure 16.: Cationic and Differential Conductivity panel.
Ordering Information
51515151 + 71180874
51515151 + 71185272
Sensor
2 x CLS15D-#1A1 for differential conductivity
Cable
Assembly
2 x CYK10-A881 (1,5m)
1 x CYK10-A881 (1,5m)
Included in cation filter (1.5 litre volume); resin included;
flow indicator in cation filter included
CM442-##M2#######
CM442-##M1#######
backplate stainless steel
backplate stainless steel
Transmitter
Option
1 x CLS15D-#1A1 for cationic conductivity
accessories please see chapter 3
Endress+Hauser
24
Modular Analytical Panel System with Sample Conditioner
2.11 CA71 analyzer panel
C-A100303-91
Pmax inlet=10bar
Tmax=45°C
Figure 17.: CA71 analyzer panel (picture shows optional 2-channel version)
Ordering Information
51515151 + 71180385
Analyzer
1-channel version: CA71##-#1#B3##
2-channel version: CA71##-#2#D3##
Optional
VA-Flow meter for sample
See chapter 2.1 for further details.
0,1 – 0,6 L/min, switching contact (Non Ex, 200 V,
1 A, 20 VA, IP65)
accessories please see chapter 3
Endress+Hauser
25
Modular Analytical Panel System with Sample Conditioner
Preliminary design!
2.12 Sodium (Na) analyzer panel C-A100303-92
Na-Analyzer
Waltron 9032
Pmax=2bar
Tmax=55°C
Figure 18.: Design draft of Sodium analyzer panel.
Ordering Information
51515151 + 71185814
Analyzer
Sodium Analyzer Waltron 9032 (included)
analysis range: 0,1ppb – 10ppm
Optional
VA-Flow meter for sample
See chapter 2.1 for further details.
0,1 – 0,6 L/min, switching contact (Non Ex, 200 V,
1 A, 20 VA, IP65)
accessories please see chapter 3
Endress+Hauser
26
Modular Analytical Panel System with Sample Conditioner
2.13 Analyzer for degassed cationic conductivity
C-A110318-80
Figure 19.: Picture and function principle of DAC analyzer
Ordering Information
Analyzer
51515151 + C-A110318-80
Degassed cationic conductivity DAC
analysis range: 0-200mS/cm, pH 7.5-10.5
250 VAC 50/60 Hz, 50VA
Options
Endress+Hauser
assembling on a stainless steel panel
1000x800x3mm for rack mounting
27
Modular Analytical Panel System with Sample Conditioner
2.14 Analyzer for degassed cationic conductivity
C-A120327-84
Figure 20.: Picture of DAC analyzer panel
Ordering Information
51515151 + 71185701
Analyzer
Degassed cationic conductivity DAC
Remark
230 VAC 50Hz; HxW=1000mm x 600mm
works only in combination with pre-conditioning
with cation filter (e.g. panel 71184222)
Options
stainless steel back plate
Endress+Hauser
28
Modular Analytical Panel System with Sample Conditioner
Accessories for Mounting Plates
3
3.1
Aluminium Profile Frame
C-A100202-50
This frame can be used when the panels shall be wall mounted.
Variable width Y from
300mm – 1200mm
Figure 21. :
Aluminium back frame for one or up to four modular panels.
Ordering Information
Remarks
Endress+Hauser
51515151 + 71180583
Back Frame made of Aluminium
Required width Y must be indicated with the order
configure rack wise!
29
Modular Analytical Panel System with Sample Conditioner
3.2
Weather Protection Roof
C-A100202-51
Variable width Y from
300mm – 500mm
Figure 22. :
Weather protection roof for transmitter (for one or max. two panels in a row).
Ordering Information
Remarks
Endress+Hauser
51515151 + 71180772
Weather Protection Roof
Required width Y must be indicated with the order
30
Modular Analytical Panel System with Sample Conditioner
3.3
Drain Gutter
C-A100202-53
Variable width Y from
300mm – 1200mm
Figure 23. :
Drain gutter for sample or cooling water outlet, drain on one side, for direct
mounting on plate or offset mounting on rack (for one or up to four
panels in a row).
Figure 24. :
Optional tray for sample container, which can be placed in the gutter.
Ordering Information
Optional
holder for sample container (stainless
steel)
Remarks
Endress+Hauser
51515151 + 71180889
Drain Gutter
W x H x D = 150mm x 60mm x 92mm
Required width Y must be indicated with the order
configure rack wise
31
Modular Analytical Panel System with Sample Conditioner
4
Identification
4.1 Name Plate
You can find the following name plate on each mounting plate:
Endress+Hauser Conducta GmbH+Co. KG
Dieselstrasse 24
70839 Gerlingen, Germany
Modification-No. C-###########
Serial Number:
Year of Construction:
Figure 25.: Name plate.
4.2
Scope of delivery
Please see the modifications in the TSP-database for the scope of delivery. In general the scope of
delivery for a typical mounting plate comprises the following components:
•
•
•
•
A Stainless Steel mounting plate
The complete SS-tubing (12 and 6 mm) and different connection parts
Cable duct
the mounting of all components like tubing, transmitter, flow assembly etc.
!
"
Endress+Hauser
Note!
All standard components like sensors, cables, transmitters etc. are not part of the
delivery scope of the mounting plate. They have to be ordered separately!
Caution!
The sensors are not installed in the flow assemblies for shipment.
32
Modular Analytical Panel System with Sample Conditioner
5
Installation
!
5.1
5.2
Note!
For installation, commissioning and operation of the SL100/200-L/H
panels, the PE02A/PE03A panels and the panels with third party products, see the separate documentation which comes with these panels and
can be found in the TSP modification database.
Incoming Acceptance, Transport, Storage
•
Make sure the packaging is undamaged!
Inform the supplier about damage to the packaging. Keep the damaged packaging
until the matter has been settled.
•
Make sure the contents are undamaged!
Inform the supplier about damage to the delivery contents. Keep the damaged
products until the matter has been settled.
•
Check that the scope of delivery is complete and agrees with your order and the
shipping documents.
•
The packaging material used to store or to transport the product must provide
shock protection and humidity protection. The original packaging offers the best
protection. Also, keep to the approved ambient conditions (see "Technical Data
Section").
•
If you have any questions, please contact your supplier or your sales centre responsible.
Installation Requirements
"
Endress+Hauser
Attention!
The mounting panels shall be wall mounted in dry rooms (Non-Ex-areas) protected from frosty ambient conditions. They can also be installed in monitoring
containers or on free standing racks.
33
Modular Analytical Panel System with Sample Conditioner
6
Connections
!
Note!
For installation, commissioning and operation of the SL100/200-L/H
panels, the PE02A/PE03A panels, the KEP02/KEP05 panels and the panels with third party products, see the separate documentation which
comes with these panels and can be found in the TSP modification database.
6.1 Electrical Connection
Please read the operating instructions of the mounted transmitter for connection the power supply and the sensors.
!
Note!
To wire the switching contacts of the optional flow meters, you can loosen the fittings of the flow meter and rotate them 90 degrees. Then you can wire the angle
plug of the flow meter. Rotate the flow meter back into the original position and
retighten the fittings again. Please assure via a pressure test that the connection is
leak-proof.
Ensure that the protective earth is connected.
"
6.2
Caution!
Before you switch on the power supply, ensure that all transmitters are correctly
connected to the power supply.
Medium Connection
"
Caution!
Before connecting the medium supply, check that all sensors are properly installed
and connected to the transmitters.
Check if cooling water and sample is available. If a sample preparation panel is
used please read the operating instruction of this panel carefully. Always turn on
the cooling water first then turn on the medium.
6.3 Connection between two or more panels
All components to connect the sample conditioning panels to a measuring panel (e.g. CA100220-70, C-A100208-50, C-A100209-50 …) are in the scope of delivery. Connect the
sample conditioning panel with fittings for 12 mm pipes to a measuring panel.
The main sample line is blocked at the right end. To connect a further panel to the right, this
blind plug must be removed.
Endress+Hauser
34
Modular Analytical Panel System with Sample Conditioner
7
Operation
!
!
Note!
For installation, commissioning and operation of the SL100/200-L/H
panels, the PE02A/PE03A panels, the KEP02/KEP05 panels and the panels with third party products, see the separate documentation which
comes with these panels and can be found in the TSP modification database.
Note!
The operation of the instruments is described in detail in the Operating Instructions of the instruments. Please see the appropriate Operating Instruction!
Please ensure the max. allowed process pressure, process temperature, flow rate
and ambient conditions comply with the requirements of the different instruments.
!
Note!
Each mounting plate has an inlet needle valve which allows to separate each
measurement line. To separate a measurement line, e.g. for maintenance, first
close the valve of the incoming measurement line.
The flow can be adjusted with the inlet needle valve. The optional flow meter allows measuring the flow and offering an electrical switch contact, which indicates
a flow below the minimum.
#
Warning!
For max. pressure and temperature conditions for the media at the sample inlet,
refer to the technical data in chapter 11!
The outlet of the measurement plates should have no counter pressure.
Endress+Hauser
35
Modular Analytical Panel System with Sample Conditioner
!
Endress+Hauser
Note!
For installation, commissioning and operation of the SL100/200-L/H
panels, the PE02A/PE03A panels, the KEP02/KEP05 panels and the panels with third party products, see the separate documentation which
comes with these panels and can be found in the TSP modification database.
36
Modular Analytical Panel System with Sample Conditioner
8
Maintenance
"
!
Endress+Hauser
Caution!
Avoid that sand, fibres and other abrasive material can enter the system!
Note!
Please follow the maintenance recommendations in the Operating Instructions of
the sensors, cables, instruments, analyzers and assemblies.
37
Modular Analytical Panel System with Sample Conditioner
9
General accessories
!
Note!
For accessories please see the Operating Instructions of the standard products.
For the mounting plates are several accessories available. Please see chapter 3.
Endress+Hauser
38
Modular Analytical Panel System with Sample Conditioner
10 Trouble shooting
!
Endress+Hauser
Note!
Please follow the trouble shooting recommendations in the Operating Instructions
of the standard products (e.g. instruments).
39
Modular Analytical Panel System with Sample Conditioner
11 Technical Data
11.1 Process Conditions
!
Note!
This chapter describes only the technical data of the mounting panels.
Concerning the technical data of the assemblies, sensors, cables, instruments and
analyzers, please refer to the operating instructions of these products.
Panel 71181119
Panel 71178791
Max. pressure sample inlet
Sample Conditioning Panel Low Pressure
34 bar
Max. sample inlet tempera- 232 °C
ture
Panel 71180834
Panel 71180881
Max. pressure sample inlet
Pressure at the sample inlet
before pressure reduction
valve
Sample inlet temperature
before cooler
Sample Conditioning Panel High Pressure
35 – 209 bar
Max. sample inlet tempera- 538 °C
ture
Pressure at the sample inlet
before pressure reduction
valve
Sample inlet temperature
before cooler
Panel 71180908
Sample Conditioning Panel High Pressure / Low
for water / condensate
Temp.
Max. pressure sample inlet
245 bar
Pressure at the sample inlet
before pressure reduction
valve
Max. sample inlet tempera- 250°C
Sample inlet temperature
ture
before cooler
Panel 71180341
Sample Conditioning Panel High Pressure / High
for steam
Temp.
Max. pressure sample inlet
245 bar
Pressure at the sample inlet
before pressure reduction
valve
Max. sample inlet tempera- 400°C / 550°C
Sample inlet temperature
ture
before cooler
Panel 71184177
Sample Conditioning Panel High Pressure / High
for water / condensate
Temp.
Max. pressure sample inlet
65 bar
Pressure at the sample inlet
before pressure reduction
valve
Max. sample inlet tempera- 200°C
Sample inlet temperature
ture
before cooler
Endress+Hauser
40
Modular Analytical Panel System with Sample Conditioner
Panel 71184184
Sample Conditioning Panel High Pressure / High
for water / condensate
Temp.
Max. pressure sample inlet
160 / 320 bar
Pressure at the sample inlet
before pressure reduction
valve
Max. sample inlet tempera- 400°C
Sample inlet temperature
ture
before cooler
Panel 71184220
Sample Conditioning Panel High Pressure / High
for water / condensate
Temp.
Max. pressure sample inlet
250 / 320 bar
Pressure at the sample inlet
before pressure reduction
valve
Max. sample inlet tempera- 550 °C / 650°C
Sample inlet temperature
ture
before cooler
Panel
71180819
71180887/ Conductivity (CLS15D) or
DO (COS21D) or pH
(CPS11D, CPS41D) panel
Max. pressure sample inlet
4 bar
Pressure at the sample inlet
Max. sample inlet tempera- 80 °C
Sample inlet temperature
ture
Max. counter pressure outlet free outlet
Panel 71179559
Conductivity
panel
Max. pressure sample inlet
4 bar
Max. sample inlet tempera- 80 °C
ture
Max. counter pressure outlet free outlet
(CLS21D)
Conductivity (CLS15D) +
DO (COS21D) or Conductivity (CLS15D) + pH
(CPS11D, CPS41D)
Max. pressure sample inlet
4 bar
Max. sample inlet tempera- 80 °C
ture
Max. counter pressure outlet free outlet
Pressure at the sample inlet
Sample inlet temperature
Panel 71181048
DO (COS21D) + pH
(CPS11D, CPS41D)
Max. pressure sample inlet
4 bar
Max. sample inlet tempera- 80 °C
ture
Max. counter pressure outlet free outlet
Pressure at the sample inlet
Sample inlet temperature
Panel 71178418
Endress+Hauser
Pressure at the sample inlet
Sample inlet temperature
41
Modular Analytical Panel System with Sample Conditioner
Panel 71184222
Cationic / diff. conductivity panel (CLS15D)
Max. pressure sample inlet
4 bar
Max. sample inlet tempera- 70 °C
ture
Max. counter pressure outlet free outlet
Pressure at the sample inlet
Sample inlet temperature
Panel
71180874
/
71185272
Max. pressure sample inlet
Max. sample inlet temperature
Max. counter pressure outlet
Cationic / diff. conductivity panel (CLS15D)
3,2 bar
60 °C
Pressure at the sample inlet
Sample inlet temperature
Panel 71180385
Max. pressure sample inlet
Max. sample inlet temperature
Max. counter pressure outlet
CA71## Analyzer panel
10 bar
40 °C
Panel 71185814
Max. pressure sample inlet
Max. sample inlet temperature
Max. counter pressure outlet
Sodium analyzer panel
2 bar
55 °C
Panel C-A110318-80
Max. pressure sample inlet
Max. sample inlet temperature
Max. counter pressure outlet
Degassed conductivity
on request
on request
Panel 71185701
Max. pressure sample inlet
Max. sample inlet temperature
Max. counter pressure outlet
Degassed conductivity
on request
on request
General information
Power Supply
Ambient temperature range
Humidity
Endress+Hauser
free outlet
Pressure at the sample inlet
Sample inlet temperature
free outlet
Pressure at the sample inlet
Sample inlet temperature
free outlet
Pressure at the sample inlet
Sample inlet temperature
free outlet
Pressure at the sample inlet
Sample inlet temperature
free outlet
Depending on transmitter and
analyzer type
+ 2 to + 40 °C
Max. 95 % relative humidity,
non-condensing
42
Modular Analytical Panel System with Sample Conditioner
Endress+Hauser
44