Buhler Technologies Zirconia oxygen analyzer BA 1000 Instruction Manual
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
Lesen Sie die Bedienungsanleitung vor dem Gebrauch des Gerätes gründlich durch, insbesondere
die Hinweise unter Gliederungspunkt 2. Andernfalls könnten Gesundheits- oder Sachschäden
auftreten. Die Bühler Technologies GmbH haftet nicht bei eigenmächtigen Änderungen des Gerätes
oder für unsachgemäßen Gebrauch.
Read this instruction carefully prior to installation and/or use. Pay attention particularly to all
advice and safety instructions to prevent injuries. Bühler Technologies GmbH cannot be held
responsible for misusing the product or unreliable function due to unauthorised modifications.
BX550001, 10/2011
Art. Nr. 9031098
Bühler Technologies GmbH, Harkortstr. 29, D-40880 Ratingen
Tel. +49 (0) 21 02 / 49 89-0, Fax. +49 (0) 21 02 / 49 89-20
Internet: www.buehler-technologies.com
Email: analyse@buehler-technologies.com
1
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
Inhaltsverzeichnis
Seite
1
Einleitung ............................................................................................................................. 4
2
Wichtige Hinweise ............................................................................................................... 4
3
Transport und Lagerungsvorschriften .............................................................................. 6
4
Aufbauen und Anschließen ................................................................................................ 6
4.1
Elektrischer Anschluss ......................................................................................................................7
4.2
Einschraubverschraubung ................................................................................................................9
4.3
Einstellung der Grenzwertkontakte beim BA 1000 D 230 G ...........................................................10
5
Betrieb und Wartung ......................................................................................................... 11
5.1
Warnhinweise ..................................................................................................................................11
5.2
Inbetriebnahme ...............................................................................................................................11
5.2.1
Grundeinstellung .........................................................................................................................12
5.2.2
Kalibrierung .................................................................................................................................12
5.3
Wartung ...........................................................................................................................................13
5.3.1
Auswechseln der Schmelzsicherung ..........................................................................................13
5.4
Besonderheiten des BA 1000 S IP 66 zum Einsatz in Kompostmieten ..........................................14
6
Instandsetzung und Entsorgung ..................................................................................... 14
6.1
Entsorgung ......................................................................................................................................14
7
7.1
7.2
7.3
2
Anhang ............................................................................................................................... 15
Ersatzteile und Zubehör ..................................................................................................................15
Gewährleistung ...............................................................................................................................15
Beigefügte Dokumente ....................................................................................................................15
BX550001, 10/2011
Art. Nr. 9031098
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
Contents
Page
1
Introduction ........................................................................................................................16
2
Important advice ................................................................................................................16
3
Transport and storing regulations ...................................................................................17
4
Installation and connection ..............................................................................................17
4.1
Electrical connection ....................................................................................................................... 17
4.2
Screw-in threaded joint ................................................................................................................... 20
4.3
Adjustment of the limit contacts of the BA 1000 D 230 G ............................................................... 21
5
Operation and maintenance..............................................................................................22
5.1
Indication of risk .............................................................................................................................. 22
5.2
Starting ............................................................................................................................................ 22
5.2.1
Basic setting ................................................................................................................................ 23
5.2.2
Calibration ................................................................................................................................... 23
5.3
Maintenance .................................................................................................................................... 24
5.3.1
Replacing the fuse ...................................................................................................................... 24
5.4
Additional hints for BA 1000 S IP66 for decomposition .................................................................. 24
6
Repair and disposal ...........................................................................................................25
6.1
Disposal........................................................................................................................................... 25
7
7.1
7.2
7.3
Appendix.............................................................................................................................26
Spare parts and accessories .......................................................................................................... 26
Warranty .......................................................................................................................................... 26
Attached documents ....................................................................................................................... 26
BX550001, 10/2011
Art. Nr. 9031098
3
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
1 Einleitung
Der BA 1000 dient der Analyse des Sauerstoffanteils in Gasgemischen. Das Gerät darf nur in einwandfreiem Zustand unter Beachtung der Bedienungsanleitung betrieben werden. Es ist nicht zum Einsatz in
zündfähigen Gasgemischen oder explosionsgefährdeter Umgebung geeignet. Bitte beachten Sie die
Angaben der Datenblätter hinsichtlich spezifischen Verwendungszwecks, vorhandener Werkstoffkombinationen sowie Druck- und Temperaturgrenzen.
Der BA 1000 S IP66 ist gedacht für die Überwachung des Sauerstoffgehaltes in Kompostmieten während
des Verrottungsprozesses. Die Sonde ist ausgestattet mit einer feuchtebeständigen Membran sowie mit
einer durch ein IP 66 Gehäuse geschützten Elektronik. Die Sondenlänge beträgt 220 mm. Die Erläuterung
der Besonderheiten des BA 1000 S IP66 finden Sie in Kapitel 5.4.
2 Wichtige Hinweise
Bitte überprüfen Sie vor Einbau des Gerätes, ob die genannten technischen Daten den Anwendungsparametern entsprechen. Überprüfen Sie ebenfalls, ob alle zum Lieferumfang gehörenden Teile vollständig
vorhanden sind.
Der Einsatz der Geräte ist nur zulässig, wenn:
- das Produkt unter den in der Bedienungs- und Installationsanleitung beschriebenen Bedingungen, dem
Einsatz gemäß Typenschild und für Anwendungen, für die es vorgesehen ist, verwendet wird.
- die im Datenblatt und der Anleitung angegebenen Grenzwerte eingehalten werden.
- Überwachungsvorrichtungen/ Schutzvorrichtung korrekt angeschlossen sind.
- die Service- und Reparaturarbeiten, die nicht in dieser Anleitung beschrieben sind, von Bühler
Technologies GmbH durchgeführt werden.
- Originalersatzteile verwendet werden.
Diese Bedienungsanleitung ist Teil des Betriebsmittels. Der Hersteller behält sich das Recht vor, die
Leistungs-, die Spezifikations- oder die Auslegungsdaten ohne Vorankündigung zu ändern. Bewahren Sie
die Anleitung für den späteren Gebrauch auf.
Begriffsbestimmungen für Warnhinweise:
HINWEIS
Signalwort für wichtige Information zum Produkt, auf die im besonderen Maße
aufmerksam gemacht werden soll.
VORSICHT
Signalwort zur Kennzeichnung einer Gefährdung mit geringem Risiko, die zu einem
Sach-schaden oder leichten bis mittelschweren Körperverletzungen führen kann,
wenn sie nicht vermieden wird.
WARNUNG
Signalwort zur Kennzeichnung einer Gefährdung mit mittlerem Risiko, die
möglicherweise Tod oder schwere Körperverletzungen zur Folge hat, wenn sie nicht
vermieden wird.
GEFAHR
Signalwort zur Kennzeichnung einer Gefährdung mit hohem Risiko, die unmittelbar
Tod oder schwere Körperverletzung zur Folge hat, wenn sie nicht vermieden wird.
4
BX550001, 10/2011
Art. Nr. 9031098
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
Warnung vor einer
allgemeinen Gefahr
Warnung vor explosionsgefährdeten Bereichen
Netzstecker ziehen
Warnung vor
elektrischer Spannung
Warnung vor heißer
Oberfläche
Atemschutz tragen
Warnung vor dem
Einatmen giftiger Gase
Gesichtsschutz
tragen
Warnung vor ätzenden
Flüssigkeiten
Handschuhe tragen
Das Gerät darf nur von Fachpersonal installiert werden, das mit den Sicherheitsanforderungen und den
Risiken vertraut ist.
Beachten Sie unbedingt die für den Einbauort relevanten Sicherheitsvorschriften und allgemein gültigen
Regeln der Technik. Beugen Sie Störungen vor und vermeiden Sie dadurch Personen- und Sachschäden.
Der für die Anlage Verantwortliche muss sicherstellen, dass:
- Sicherheitshinweise und Betriebsanleitungen verfügbar sind und eingehalten werden,
- Unfallverhütungsvorschriften der Berufsgenossenschaften beachtet werden; in Deutschland:
Allgemeine Vorschriften” (VBG 1) und “Elektrische Anlagen und Betriebsmittel (VBG 4)”,
- die zulässigen Daten und Einsatzbedingungen eingehalten werden,
- Schutzeinrichtungen verwendet werden und vorgeschriebene Wartungsarbeiten durchgeführt werden,
- bei der Entsorgung die gesetzlichen Regelungen beachtet werden.
Wartung, Reparatur:
- Reparaturen an den Betriebsmitteln dürfen nur von Bühler autorisiertem Personal ausgeführt werden.
- Nur Umbau-, Wartungs- oder Montagearbeiten ausführen, die in dieser Bedienungs- und
Installationsanleitung beschrieben sind.
- Nur Original-Ersatzteile verwenden.
Bei Durchführung von Wartungsarbeiten jeglicher Art müssen die relevanten Sicherheits- und
Betriebsbestimmungen beachtet werden.
!
GEFAHR
Elektrische Spannung
Gefahr eines elektrischen Schlages.
Trennen Sie das Gerät vor Beginn der Wartungsarbeiten vom Netz. Sichern
Sie das Gerät gegen unbeabsichtigtes Wiedereinschalten.
Anschluss und Wartung dürfen nur von geschultem Fachpersonal
vorgenommen werden. Achten Sie auf die korrekte Spannungsversorgung!
BX550001, 10/2011
Art. Nr. 9031098
5
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
!
GEFAHR
Giftige, ätzende Gase
Messgas kann gesundheitsgefährdend sein.
Sorgen Sie ggf. für eine sichere Ableitung des Gases.
Stellen Sie vor Beginn der Wartungsarbeiten die Gaszufuhr ab und sichern
Sie sie gegen unbeabsichtigtes Aufdrehen.
Schützen Sie sich bei der Wartung vor giftigen / ätzenden Gasen. Tragen Sie
die entsprechende Schutzausrüstung.
!
GEFAHR
Explosionsgefahr bei Verwendung in Explosionsgefährdeten Bereichen
Das Betriebsmittel ist nicht für den Einsatz in explosionsgefährdeten
Bereichen geeignet.
Durch das Gerät dürfen keine zündfähigen oder explosiven Gasgemische
geleitet werden.
3 Transport und Lagerungsvorschriften
Die Betriebsmittel nur in der Originalverpackung oder einem geeigneten Ersatz transportieren. Bei längerer
Nichtbenutzung sind sie gegen Feuchtigkeit und Wärme zu schützen.
Die Betriebsmittel müssen in einem überdachten, trockenen, vibrations- und staubfreien Raum bei einer
Temperatur von –20°C bis +50°C aufbewahrt werden.
4 Aufbauen und Anschließen
Das Gerät darf nur von Fachpersonal installiert werden, das mit den Sicherheitsanforderungen und den
Risiken vertraut ist.
Beachten Sie unbedingt die für den Einbauort relevanten Sicherheitsvorschriften und allgemein gültigen
Regeln der Technik. Beugen Sie Störungen vor und vermeiden Sie dadurch Personen- und Sachschäden.
Wo immer möglich, sollte der O2-Sensor nach unten zeigend eingebaut werden.
Dies muss immer dann erfolgen, wenn im Prozess ein sehr hoher Feuchteanteil
vorhanden ist. Kommt es zur Kondensatbildung, kann der Einbau eines zusätzlichen
Schutzrohres erforderlich sein. In trockenen Applikationen kann das Gerät auch
waagerecht installiert werden.
Der
Montageort
sollte
jederzeit
zugänglich
und
einsehbar
Umgebungsbedingungen sollen die zulässigen Werte nicht überschreiten.
sein.
Die
Bei Montage im Freien sollte der BA 1000 vor direkten Witterungseinwirkungen
geschützt werden. Der O2-Sensor darf von Wasser oder Spritzwasser nicht erreicht
werden. Die Montage in Feuchträumen ist unzulässig.
6
BX550001, 10/2011
Art. Nr. 9031098
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
4.1 Elektrischer Anschluss
WARNUNG
Der Anschluss darf nur von geschultem Fachpersonal vorgenommen werden.
Auf korrekte Spannungsversorgung achten. Örtliche Vorschriften beachten.
HINWEIS
!
Signalleitungen und Starkstromkabel nicht nebeneinander verlegen um die
Gefahr von Störeinstrahlungen zu vermeiden. Ggf. abgeschirmtes Kabel
verwenden.
Beachten Sie zudem die nachfolgende Abbildung, um Fehler bei der elektrischen Verdrahtung zu
vermeiden.
Buchse J1: Stromversorgung
Buchse J3: Messfühlerleitung
grün -gelb*
2
PE
Nr. 2*
3
GND
weiß*
GND
grün*
0-10 V
IMP
4
braun*
4-20 mA
Nr. 1*
1
+ 24 V
GND
2 5
3
1
*Die Kabelzeichnungen gelten für das mitgelieferte Verbindungskabel Art.-Nr. 55012995
Netzteil BA 1000 N:
1
2
3
4
5
6
+
-
PE
L1
N
PE
⏐
⏐
⏐
⏐
↓
↓
24V DC
⏐
⏐
⏐
⏐
⏐
⏐
↑
↑
↑
Netz 230V AC
48 – 62 Hz
Digitalanzeige BA 1000 D 24:
1
2
3
4
+
-
+
⏐
⏐
⏐
⏐
⏐
⏐
↑
↑
↓
4 – 20 mA
7
8
-
+
-
PE
⏐
⏐
↓
⏐
⏐
↑
⏐
⏐
↑
5 Abschirmung
6
BX550001, 10/2011
Art. Nr. 9031098
7
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
Digitalanzeige BA 1000 D 230
7
8
9
10
11
-
+
-
PE
L1
N
PE
⏐
⏐
↓
⏐
⏐
⏐
⏐
↓
↓
24V DC
2
3
4
+
-
+
⏐
⏐
⏐
⏐
⏐
⏐
↑
↑
↓
4 – 20 mA
5 Abschirmung
6
1
⏐
⏐
⏐
⏐
⏐
⏐
↑
↑
↑
Netz 230V AC
48 – 62 Hz
1
2
3
4
+
-
+
-
⏐
⏐
⏐
⏐
⏐
⏐
↑
↑
↓
4 – 20 mA
⏐
⏐
↓
5 Abschirmung
Digitalanzeige BA 1000 D 230 G
6
7
8
+
-
PE
⏐
⏐
⏐
⏐
↓
↓
24 V DC
⎡
⎢
⎢
9
\
\
10
⎤
⎥
⎥
11
⎡
⎢
⎢
12
\
\
13
Grenzwertkontakte
⎤
⎥
⎥
14
15
16
17
L1
N
PE
⏐
⏐
⏐
⏐
⏐
⏐
↑
↑
↑
Netz 230V AC
48 – 62 Hz
Beim Anschluss des 4-20 mA- Ausgangs muss in den Digitalanzeigen die Brücke zwischen Klemme 3 und
4 entfernt werden.
Beachten Sie die erforderlichen Leitungsquerschnitte in Abhängigkeit von der Kabellänge:
Kabellänge Leitungsquerschnitt
30 m
0,25 mm2
60 m
0,5 mm2
100 m
0,75 mm2
Weiterhin ist folgender Einschaltstrom zu beachten.
8
BX550001, 10/2011
Art. Nr. 9031098
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
A
Einschaltstrom
4,4
2,6
0,6
0
15ms
70
ms
4.2 Einschraubverschraubung
Es stehen 2 Einschraubverschraubungen für den Einbau des O2 Sensors zur Auswahl
•
1/2“ NPT Einschraubverschraubung mit PTFE Klemmringen zur stufenlosen
Längeneinstellung des O2 Sensors
Artikelnummer
55012993
•
1/2“ NPT Einschraubverschraubung mit Edelstahl Klemmringen
Artikelnummer
55012994
Bei dieser Ausführung ist nach dem Festziehen der Mutter keine nachträgliche Längeneinstellung des O2 Sensors möglich. (Der Schneidring wird im Sondenrohr eingeschnitten.)
∅12
NPT - 1/2"
Montage
BX550001, 10/2011
Art. Nr. 9031098
9
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
4.3 Einstellung der Grenzwertkontakte beim BA 1000 D 230 G
Grau hinterlegte Tasten dauernd gedrückt halten. Im Anzeigemodus haben die Tasten folgende
Funktionen:
GW1
Anzeige Schaltausgang 1
GW2
Anzeige Schaltausgang 2
P
Anzeige Geräteversion
<
Anzeige Minimalwert
>
Anzeige Maximalwert
<
+
P
lösche Minimalwert
>
+
P
lösche Maximalwert
P
+2x
GW1
Änderungsmodus Schaltausgang 1
P
+2x
GW2
Änderungsmodus Schaltausgang 2
P
+
GW1
+2x
GW2 Änderungsmodus Grundeinstellungen
Tastenfunktionen im Änderungsmodus
GW2
blättert vorwärts, zum nächsten Einstellpunkt
GW1
blättert rückwärts, zum vorherigen Einstellpunkt
<
wählt die zu ändernde Stelle ( blinkend ) an
>
zählt die blinkenden Ziffern hoch
P
übernimmt Einstellungen, Ende Änderungsmodus
Erfolgt zwei Minuten lang keine Eingabe, so schaltet das Gerät ohne Übernahme der Änderungen zurück
in den Anzeigemodus.
Änderungsmodus Schaltausgänge
P
+2x GW1
Änderungsmodus Schaltausgang 1
blättern
GW2
GW1
P
+
GW1
Anzeige
Einstellpunkt
Bereich
S1. S
S1. H
S1. An
S1. Ab
Schaltpunkt
Schalthysterese
Abzugsverzögerung ( in s )
Abfallverzögerung ( in s )
-19999...+19999
1...19999 Digit
0...999,9s
0...999,9s
+2x
GW2 Änderungsmodus Grundeinstellungen
Einstellpunkte siehe oben
10
BX550001, 10/2011
Art. Nr. 9031098
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
5 Betrieb und Wartung
5.1 Warnhinweise
Das Gerät darf nur von Fachpersonal installiert
Sicherheitsanforderungen und den Risiken vertraut ist.
und
gewartet
werden,
das
mit
den
- Das Gerät darf nicht außerhalb seiner Spezifikationen betrieben werden.
- Reparaturen an den Betriebsmitteln dürfen nur von Bühler autorisiertem Personal ausgeführt werden.
- Führen Sie nur Umbau-, Wartungs- oder Montagearbeiten aus, die in dieser Bedienungs- und
Installationsanleitung beschrieben sind.
- Verwenden Sie nur Original-Ersatzteile.
- Beachten Sie bei der Durchführung von Wartungsarbeiten jeglicher Art die relevanten Sicherheits- und
Betriebsbestimmungen.
GEFAHR
Elektrische Spannung
Gefahr eines elektrischen Schlages.
Trennen Sie das Gerät bei allen Wartungsarbeiten vom Netz. Sichern Sie das
Gerät gegen unbeabsichtigtes Wiedereinschalten.
Der Anschluss darf nur von geschultem Fachpersonal vorgenommen werden.
Achten Sie auf die korrekte Spannungsversorgung!
!
VORSICHT
Spannungsverlust
Beschädigung des Gerätes durch fehlende Versorgungsspannung
Der BA 1000 muss im eingebauten Zustand immer mit Spannung versorgt
werden, um den Sensor auf Betriebstemperatur zu halten. Dies ist erforderlich,
damit es am Sensor auf keinen Fall zu Kondensatbildung kommt.
5.2 Inbetriebnahme
Bei Einsatz in Kompostmieten siehe auch Kapitel 5.4
Vergewissern Sie sich vor der Inbetriebnahme, ob der Prozess unter Normalbedingungen läuft, alle
Installationshinweise beachtet wurden und die korrekte Spannungsversorgung vorliegt.
Überprüfen Sie weiterhin, ob alle Verbindungen korrekt ausgeführt sind. Lediglich der Deckel am O2Sensor kann geöffnet bleiben, damit die Grundeinstellung überprüft bzw. vorgenommen werden kann.
BX550001, 10/2011
Art. Nr. 9031098
11
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
5.2.1 Grundeinstellung
An jedem BA 1000 wird vor Auslieferung ab Werk die Grundeinstellung vorgenommen. Diese kann vor Ort
wie folgt überprüft werden:
¾
Schließen Sie an die Buchse J3 ein Amperemeter oder eine Kalibriereinheit BA 1000 KA an.
¾
Ziehen Sie den Sensor aus dem Prozess und halten Sie ihn für einige Minuten in die
Umgebungsluft.
Die Kalibriereinheit sollte bei Normalbedingungen (1013 mbar / 20°C) 20,9 bzw. 21 Vol% O2 anzeigen, das
Amperemeter 17,44 mA.
Dabei wird vorausgesetzt, dass die Umgebungsluft 20,9 Vol% O2 hat. Wird die Prüfung unter
Umgebungsbedingungen vorgenommen, in denen der Normal O2-Gehalt zweifelhaft ist, muss ggf. Prüfgas
verwendet werden. Die Sauerstoffanzeige ist abhängig von den Umgebungsbedingungen (Luftdruck,
Temperatur).
Bei der Überprüfung der Grundeinstellung nach längerem Betrieb des Analysators kann es erforderlich
sein, dass dieser erst längere Zeit an Umgebungsluft betrieben wird, bevor die Grundeinstellung wieder
erreicht wird. Dies hängt damit zusammen, dass aus dem Prozess stammende Gaskomponenten zunächst
aus der Messzelle ausdiffundieren müssen.
5.2.2 Kalibrierung
Der Endwert kann mit dem Span- Potentiometer eingestellt werden. Hierzu muss der Gehäusedeckel am
O2 Sensor abgenommen werden.
Sicherung/
Fuse
Endwert/
Span
A000106X
Sicherung/
Fuse
Endwert/
Span
Abb.1 Innenansicht Standardgehäuse (links) und IP 66 – Gehäuse (rechts)
!
VORSICHT
Die Potentiometer Uref und Zero dürfen nicht verstellt werden!
12
BX550001, 10/2011
Art. Nr. 9031098
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
Das Potentiometer Span dient zum Abgleich der Sauerstoffkonzentration.
20,9 Vol.-% O2 = 17,44 mA = 8,4V (bei 1013mbar/20°C)
Span
Zero
Uref
5.3 Wartung
Der BA 1000 arbeitet wartungsfrei. Je nach Anwendung empfiehlt sich von Zeit zu Zeit die Überprüfung der
Grundeinstellung. Dabei ist zu beachten, dass die Druckverhältnisse innerhalb des Prozesses und in der
Umgebungsluft von einander abweichen können. Dies führt zu Abweichungen in den Ausgabewerten.
Um die Beeinflussung des Sensors der Messzelle durch Kondensat zu verhindern, empfehlen wir den
Sensor aus dem Prozess zu ziehen.
!
WARNUNG
Heiße Oberfläche
Verbrennungsgefahr bei Berührung der Sondenspitze
Die Sondenspitze ist heiß. Berühren Sie die Sondenspitze nicht und tragen
Sie Schutzhandschuhe.
Trennen Sie ihn danach vom Netz, wenn Sie sichergestellt haben, dass kein Kondensat in die Messzelle
gelangen kann.
5.3.1 Auswechseln der Schmelzsicherung
Die Sicherung befindet sich im Gehäuse des O2-Sensors (siehe Abb. 1).
¾
Sensor aus dem Prozess ziehen.
¾
Sicherstellen, dass kein Kondensat in die Messzelle eindringen kann.
¾
Gerät vom Netz trennen und gegen wiedereinschalten sichern und Gehäuse öffnen.
¾
Schraubkappe von der Sicherung abnehmen.
¾
Schmelzsicherung 0,8 A träge ersetzen.
¾
Schraubkappe aufsetzen und wieder anziehen, Gehäusedeckel schließen, Gerät wieder ans Netz
anschließen.
BX550001, 10/2011
Art. Nr. 9031098
13
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
5.4 Besonderheiten des BA 1000 S IP 66 zum Einsatz in
Kompostmieten
- Der Analysator darf nur in Verbindung mit einer Bodenlanze zur Überwachung von Kompostmieten
eingesetzt werden.
- Die Kalibrierung des Gerätes darf nur in trockener Umgebung durchgeführt werden, um ein Eindringen
von Feuchtigkeit in die Elektronik zu vermeiden.
- Sollte dies offensichtlich nicht der Fall sein und es zu starken Beschädigungen der Elektronik
kommen verfällt der Garantieanspruch.
!
HINWEIS
Die Messzelle, einschließlich der Heizung, ist ein Verschleißteil und von der Gewährleistung ausgenommen. Ansonsten gelten unsere allgemeinen Geschäftsbedingungen.
- Die Kalibrierung verläuft wie unter Punkt 4.1.2 beschrieben.
- Das Elektronikgehäuse ist nach der Kalibrierung wieder absolut dicht zu verschließen. (Ein
Dichtungswechsel am Elektronikgehäuse ist nach 4-5-maligem Kalibrieren zu empfehlen)
- Sollte Feuchtigkeit in die Membran des Sensors dringen kann dieser nach Abtrocknen an der Luft
wiederverwendet werden.
- Eine Reinigung der Sonde mit einem Dampfstrahler zerstört das Gerät und ist nicht gestattet.
- Hohe Konzentrationen an Schwermetallen und Schwefelverbindungen, sowie ein Ammoniakgehalt von
> 2000 ppm beeinflussen die Lebensdauer der Messzelle.
- Während des Betriebes in der Kompostmiete, darf das Gerät nicht von der Spannungsquelle
abgeschaltet werden. (ÆAbkühlung der Heizung hat evtl. Kondensateinbrüche zur Folge.)
6 Instandsetzung und Entsorgung
Sollte ein Fehler beim Betrieb auftreten, muss das Gerät durch den Hersteller überprüft werden.
Sollten Sie Fragen haben, wenden Sie sich bitte an unseren Service
Tel.: +49-(0)2102-498955 oder Ihre zuständige Vertretung.
Bitte senden Sie das Gerät in geeigneter Verpackung an:
Bühler Technologies GmbH
- Reparatur/Service Harkortstraße 29
40880 Ratingen
Deutschland
Bringen Sie zusätzlich die Dekontaminierungserklärung ausgefüllt und unterschrieben an der Verpackung
an. Ansonsten ist eine Bearbeitung Ihres Reparaturauftrages nicht möglich! Das Formular kann per E-Mail
angefordert werden: service@buehler-technologies.com.
6.1 Entsorgung
Bei der Entsorgung sind die gesetzlichen und Normenregelungen im Anwenderland zu beachten.
14
BX550001, 10/2011
Art. Nr. 9031098
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
7 Anhang
7.1 Ersatzteile und Zubehör
Art. Nr.
Bezeichnung
5501599
Netzteil BA 1000 N, 115/ 230 V, 50/ 60 Hz, 24 V DC
5501699
Digitalanzeige 3 ½ stellig, BA 1000 D24, 24 V DC
5501799
Digitalanzeige 3 ½ stellig, incl. Netzteil, BA 1000 D230, 24 V DC
5501899
Digitalanzeige 4 ½ stellig, incl. Netzteil, BA 1000 D230G, 24 V DC
55012992
Kalibrieranzeige BA 1000 KA
55012993
Einschraubverschraubung ½“ NPT/ PTFE
55012994
Einschraubverschraubung ½“ NPT/ SS
9146100094
Stecker (Versorgung) J1-09-5106-15-03
9146100095
Stecker (Signalausgang) J3-09-5117-15-05
55013991
Sonde 380 mm
55014991
Sonde 780 mm
5501299
Elektronik Standardsonde
5502399
Bodensonde zum Einsatz in Kompostmieten
5502009953
Ersatzsonde 220 mm
550200995110 Elektronik (Kompostsonde) IP 66
7.2 Gewährleistung
Die Messzelle, einschließlich der Heizung, ist ein Verschleißteil und von der Gewährleistung
ausgenommen!
Ansonsten gelten unsere allgemeinen Geschäftsbedingungen.
7.3
Beigefügte Dokumente
- Datenblatt BA 1000.
DD/DE/DA 550001
- Konformitätserklärung:
KX 550007
- Dekontaminierungserklärung
BX550001, 10/2011
Art. Nr. 9031098
15
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
1 Introduction
The BA 1000 is for analysing the proportion of oxygen in gaseous mixtures. The unit may only be operated
when it is in perfect condition and the operating instructions are followed. It is not suitable for use in
ignitable gas mixtures or in surroundings endangered by explosion. Please take note of the information on
the data sheets with regard to specific purpose and material combinations, as well as pressure and
temperature limits.
The BA 1000 S IP 66 is designed for monitoring oxygen in the decomposition during the rotting process.
The probe is equipped with a damp-steady diaphragm as well as with IP 66-protected electronics. The
probe length is 220 mm. For additional description of the BA 1000 S IP 66 see chapter 5.4.
2 Important advice
Please check before installation of the device that the technical data matches the application parameters.
Also, check that the delivery is complete.
Operation of the device is only valid if
- the product is used under the conditions described in the installation- and operation instruction, the
intended application according to the type plate and the intended use,
- the performance limits given in the datasheets and in the installation- and operation instruction are
obeyed,
- monitoring devices and safety devices are installed properly,
- service and repair is carried out by Bühler Technologies GmbH, unless described in this manual,
- only original spare parts are used.
This manual is part of the equipment. The manufacturer keeps the right to modify specifications without
advanced notice. Keep this manual for later use.
Definitions for warnings:
NOTE
Signal word for important information to the product.
CAUTION
Signal word for a hazardous situation with low risk, resulting in damaged to the device
or the property or minor or medium injuries if not avoided.
WARNING
Signal word for a hazardous situation with medium risk, possibly resulting in severe
injuries or death if not avoided.
DANGER
Signal word for an imminent danger with high risk, resulting in severe injuries or death
if not avoided.
16
Warning against
hazardous situation
Warning against possible
explosive atmospheres
disconnect from
mains
Warning against
electrical voltage
Warning against hot
surface
wear respirator
Warning against
respiration of toxic gases
wear face
protection
Warning against acid
and corrosive
substances
wear gloves
BX550001, 10/2011
Art. Nr. 9031098
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
3 Transport and storing regulations
The equipment should be only transported in the original case or in appropriate packing.
Protect the equipment against heat and humidity.
The level switch must be stored in roofed, dry, vibration- und dust free room. Temperature should be
between –20°C and +50°C (-4°F and 122°F).
4 Installation and connection
Take all necessary precautions and follow all safety regulations where you are installing the unit. The
installation and connection may only be done by authorized members of staff. Check the power supply
before connection!
Wherever possible, the O2 sensor should be installed pointing down.
This must always be done when a high proportion of moisture is in the process. If
condensates should form, the installation of an additional protective pipe can be
necessary. In dry applications, the unit can also be installed horizontally.
The place of mounting should be accessible and visible at all times. The surroundings
should be suitable.
If mounted outdoors, the BA 1000 should be protected against the direct effects of
weather. The O2 sensor must not be exposed to water or water spray. Mounting in damp
rooms is not permissible.
4.1 Electrical connection
WARNING
The device must be installed by trained staff only.
Regard the correct mains voltage as given on the type plate.
NOTE
!
To avoid the danger of spurious irradiation, do not lay signal lines and highvoltage cables next to one another. If appropriate, use shielded cables.
Please note also the following illustration to avoid errors in wiring.
Jack J1: Power supply
green -yellow*
2
PE
Nr. 2*
3
GND
Jack J3: Sensor line
green*
0-10 V
IMP
Nr. 1*
1
+ 24 V
white* brown*
GND 4-20 mA
4
GND
2 5
1
BX550001, 10/2011
3
Art. Nr. 9031098
17
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
* The cable drawings refer to the supplied connection cable, part no. 55012995
BA 1000 N power pack:
1
2
3
4
5
6
+
-
PE
L1
N
PE
⏐
⏐
⏐
⏐
↓
↓
24V DC
⏐
⏐
⏐
⏐
⏐
⏐
↑
↑
↑
Mains 230V AC
48 – 62 Hz
BA 1000 D 24 digital display:
7
8
-
+
-
PE
⏐
⏐
↓
⏐
⏐
↑
⏐
⏐
↑
6
7
8
9
10
11
-
PE
L1
N
PE
2
3
4
+
-
+
⏐
⏐
⏐
⏐
⏐
⏐
↑
↑
↓
4 – 20 mA
5 shielding
6
1
BA 1000 D 230 digital display:
18
2
3
4
+
-
+
-
+
⏐
⏐
⏐
⏐
⏐
⏐
↑
↑
↓
4 – 20 mA
⏐
⏐
↓
⏐
⏐
⏐
⏐
↓
↓
24V DC
⏐
⏐
⏐
⏐
⏐
⏐
↑
↑
↑
Mains 230V AC
48 – 62 Hz
BX550001, 10/2011
Art. Nr. 9031098
5 Shielding
1
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
1
2
3
4
+
-
+
-
⏐
⏐
⏐
⏐
⏐
⏐
↑
↑
↓
4 – 20 mA
⏐
⏐
↓
5 Shielding
BA 1000 D 230 G digital display:
⎡
⎢
⎢
9
6
7
8
+
-
PE
\
\
10
⏐
⏐
⏐
⏐
↓
↓
24 V DC
⎤
⎥
⎥
11
⎡
⎢
⎢
12
\
\
13
Limit contacts
⎤
⎥
⎥
14
15
16
17
L1
N
PE
⏐
⏐
⏐
⏐
⏐
⏐
↑
↑
↑
Mains 230V AC
48 – 62 Hz
When the 4-20 mA- output is connected, the bridge between terminals 3 and 4 in the digital displays must
be removed.
Please note the required wire diameters depending on the cable length:
Cable length
30 m
60 m
100 m
Cross-section
(approx. 98.5 ft)
0.25 mm2
(AWG 23)
(approx. 197 ft)
0.5 mm
2
(AWG 21)
0.75 mm
2
(AWG 19)
(approx. 328 ft)
In addition, note the following starting current.
A
starting current
4,4
2,6
0,6
0
15ms
70
ms
BX550001, 10/2011
Art. Nr. 9031098
19
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
4.2 Screw-in threaded joint
There are two screw-in threaded joints to choose from for the installation of the sensor
•
1/2“ NPT screw-in threaded joint with PTFE clamp collars for gradual adjustment of the length
of the O2 sensor; part number 55012993
•
1/2“ NPT screw-in threaded joint with special steel clamp collars; part number 55012994
This model does not allow the length to be adjusted after the nut has been tightened. (The
cutting ring is cut into the tube of the probe)
Assembly
NPT - 1/2"
∅12
20
BX550001, 10/2011
Art. Nr. 9031098
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
4.3 Adjustment of the limit contacts of the BA 1000 D 230 G
Press the gray keys and hold them down. In display mode, the keys have the following functions:
GW1
Switch output display 1
GW2
Switch output display 2
P
Unit version display
<
Minimum value display
>
Maximum value display
<
+
P
Delete minimum value
>
+
P
Delete maximum value
P
+2x
GW1
Modification mode for switch output 1
P
+2x
GW2
Modification mode for switch output 2
P
+
GW1
+2x
GW2 Modification mode basic settings
Key functions in modification mode
GW2
scrolls forward to next index point
GW1
scrolls back to previous index point
<
selects the place to be modified ( blinks )
>
counts up the blinking numbers
P
takes over settings, end modification mode
If nothing is entered for two minutes, the unit will switch back to display mode without taking over the
modifications.
Display mode switch outputs
P
+2x GW1
scroll
GW2
GW1
P
+2x
Modification mode for switch output 1
Display
Index point
Range
S1. S
S1. H
S1. An
S1. Ab
switching point
switching hysteresis
discharge delay ( in s )
drop-out delay ( in s )
-19999...+19999
1...19999 Digit
0...999,9s
0...999,9s
GW2
Modification mode for switch output 2
Index points see above
BX550001, 10/2011
Art. Nr. 9031098
21
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
5 Operation and maintenance
5.1 Indication of risk
Installation and maintenance of the device shall be performed by trained staff only, familiar with the safety
requirements and risks.
-
The device must be operated within its specifications.
All repairs must be carried out by Bühler authorised personnel only.
Only perform modifications, servicing, or mounting described in this manual.
Only use original spare parts.
Regard all relevant safety regulations and internal operating instructions during maintenance.
DANGER
Electrical voltage
Electrocution hazard.
Before any manipulation on the device, disconnect the electrical equipment
from mains power supply. Make sure that the electrical equipment cannot be
reconnected during repair or maintenance.
The wiring must be done by trained staff only. Regard the correct mains
voltage.
!
CAUTION
Voltage drop
Damage of the device due to voltage drop
The BA 1000 must always be supplied with power when it is installed in order
to keep the sensors at operating temperature. This is necessary to ensure that
no condensate forms on the sensor.
5.2 Starting
For use in compost windrows see also chapter 5.4 for special hints.
Before starting, make sure that the process will run under normal conditions, and that all installation
instructions have been observed and the correct voltage supply is available.
Also, check that all connections have been correctly made. The cover on the O2-sensor alone can remain
open so that the basic setting can be checked or set.
22
BX550001, 10/2011
Art. Nr. 9031098
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
5.2.1
Basic setting
The basic setting has been set on every BA 1000 before it leaves the factory. This can be checked on site
as follows:
¾
Hook up an ampere meter or BA 1000 KA calibration unit to the J3 socket.
¾
Pull the sensor out of the process and hold it in the air for a few minutes.
The calibration unit should now display 20.9 or 21Vol% O2, the ampere meter 17.44 mA. (Normal
conditions assumed: 1013 mbar / 20°C (14.69 psi / 68°F)).
This implies that the ambient air contains 20.9Vol% O2. In certain cases, a calibration gas must be used if
the O2-content of ambient air cannot be guaranteed. The oxygen value depends on ambient conditions
(barometric pressure / temperature).
When the basic setting is checked after the analyser has been in operation for a long period, it may be
necessary first to operate it for some time in ambient air before the basic setting is reached again. This is
related to the fact that the gas components from the process first have to diffuse out of the measuring cell.
5.2.2
Calibration
The final value can be set with the span potentiometer. To do so, remove the case cover of the O2 sensor.
Sicherung/
Fuse
Endwert/
Span
Sicherung/
Fuse
Endwert/
Span
A000106X
Fig.1 Interior view of the Standard housing (left) and the IP 66 model (right)
!
CAUTION
The Uref and Zero potentiometers must not be adjusted!
The span potentiometer is for balancing the oxygen concentration.
Span
Zero
Uref
20.9 vol.- % O2 = 17.44 mA = 8.4V (normal conditions 1013 mbar /
20°C)
BX550001, 10/2011
Art. Nr. 9031098
23
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
5.3 Maintenance
The BA 1000 requires no maintenance. Depending on the application, a check of the basic setting is
advisable from time to time. In doing so, note that the pressure conditions in the process and in the
surrounding air can deviate from each other. This leads to deviations in the output values.
!
WARNING
Hot surface
Burning hazard when touching the probe.
The tip of the probe is hot. Do not touch the probe and wear protection gloves.
In order to prevent condensate from affecting the sensor on the measuring cell, we recommend pulling the
sensor out of the process. Thereafter, unplug it when you have made sure that condensate can get into the
measuring cell.
5.3.1
Replacing the fuse
¾
Move the sensor out of the process.
¾
Make sure that no condensate can form inside the measuring cell.
¾
Disconnect the device from mains supply, make sure that it cannot be switched on unintentionally,
and open the cover.
¾
Take screw-on cap off the fuse.
¾
• Replace 0.8 A slow-acting fuse.
¾
• Put screw-on cap back on and tighten it.
¾
Close the cover and reconnect the device to mains supply.
5.4 Additional hints for BA 1000 S IP66 for decomposition
- The analyser may be used only in connection with a soil lance for the monitoring of the decomposition.
- The calibration of the equipment should be done only in dry environment, in order to avoid a penetration
of humidity into electronics.
- In case of not following these instructions the electronic may be damaged and therefore you
loose your warranty.
!
NOTE
The measuring cell, including the heating unit, is subject to wear and tear and
not covered by the warranty. See as well the general warranty regulations.
- The calibration runs as described in chapter 4.1.2.
- The electronics housing should be closed after every calibration again very tight. (We recommend
replacing the seal every 4-5 calibrations.)
- Should humidity penetrate into the diaphragm of the sensor, it could be reused after drying with air.
- Cleaning the probe using a steam jet destroys the equipment and is not permitted.
- High concentrations of heavy metals and sulphur compounds, as well as an ammonia concentration
> 2000 ppm will affect the life span of the measuring probe.
- During the measuring process in the compost, the equipment may not be switched off from the power
supply. (ÆCooling down of the heater could effect that condensate can get into the measuring cell.)
24
BX550001, 10/2011
Art. Nr. 9031098
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
6 Repair and disposal
If the device shows irregularities, the manufacturer must check the device.
If you need help or more information
call +49(0)2102-498955 or your local agent.
Please hold the data of the type plate ready.
Please ship the device with suitable packing to
Bühler Technologies GmbH
- Service Harkortstraße 29
40880 Ratingen
Germany
In Addition, attach the filled in and signed Declaration of Decontamination status to the packing. Otherwise,
your repair order cannot be processed! The form can be requested by e-mail to service@buehlertechnologies.com.
6.1 Disposal
Regard the local regulations for disposal of electric and electronic equipment.
BX550001, 10/2011
Art. Nr. 9031098
25
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Zirkonium Sauerstoffanalysator BA 1000
Circonia oxygen analyzer BA 1000
7 Appendix
7.1 Spare parts and accessories
Part No.
Designation
5501599
BA 1000 N power pack, 115/ 230 V, 50/ 60 Hz, 24 V DC
5501699
Digital display 3 ½ places, BA 1000 D24, 24 V DC
5501799
Digital display 3 ½ places, including power pack, BA 1000 D230, 24 V DC
5501899
Digital display 4 ½ places, including power pack, BA 1000 D230G, 24 V DC
55012992
BA 1000 KA calibration display
55012993
Screw-in threaded joint ½“ NPT/ PTFE
55012994
Screw-in threaded joint ½“ NPT/ SS
9146100094
Plug (power supply) J1-09-5106-15-03
9146100095
Plug (signal output) J3-09-5117-15-05
55013991
Probe 380 mm
55014991
Probe 780 mm
5501299
Electronic standard probe
5502399
Probe for use in compost
5502009953
Probe 220 mm
550200995110 Electronic (compost probe) IP 66
7.2 Warranty
The measuring cell, including the heating unit, is subject to wear and tear and not covered by the warranty!
See as well the general warranty regulations.
7.3 Attached documents
- Data sheet BA 1000:
DD/DE/DA 550001
- Declaration of conformity: KX 550007
- Declaration of Contamination status
26
BX550001, 10/2011
Art. Nr. 9031098
Zirkonium-Sauerstoffanalysator
BA 1000
Beschreibung
Der BA 1000 wurde für den Einsatz als Insitu-Sauerstoffsensor
entwickelt. Das Messsystem des Sensors ist selbstüberwachend und daher für die Kontrolle von Feuerungsanlagen bestens geeignet.
Funktion der Sauerstoffsonde
Der dynamische Sauerstoffsensor ermöglicht es, Sauerstoffmessungen ohne Referenzgas durchzuführen.
Der Sensor besteht aus zwei identischen Zirkondioxidscheiben, welche einen Platinring abdecken. Es entsteht eine
dichte Kammer (P2), in ihr befindet sich ein Gas mit
unbekanntem Sauerstoff-Partialdruck.
Eine Scheibe ist mit einer umkehrbaren Stromquelle verbunden, sie dient als elektrochemische Sauerstoffpumpe.
Die andere Scheibe ermittelt den O2-Partialdruck über die
erzeugte Nernst`sche Spannung.
Sensor
Glas / glass
werden wieder Sauerstoffionen in die Kammer gepumpt. Der
Partialdruck P2 steigt an bis er wieder eine bestimmte Spannung
erreicht hat, jetzt ändert sich die Stromrichtung erneut. Dieser
Vorgang wiederholt sich periodisch. Die Periodenzeit ist linear
zum Sauerstoffpartialdruck.
Dieses Prinzip ermöglicht es, den Sensor ständig zu
überwachen; der Sensor ist "fehlersicher"!
Die Kalibrierung der Sonde erfolgt an atmosphärischer Luft.
Dabei sind keine zusätzlichen Kalibriergase erforderlich. Ein
optionaler Vorfilter ermöglicht den Einsatz auch bei höherer
Staubbelastung.
Anwendungsgebiete:
Industrielle Heizanlagen, Industrielle Verbrennungsanlagen,
Industrielle Prozesssteuerungsanlagen, Bio-Technik,
Landwirtschaft (Silo), Medizin.
Eigenschaften
GND
§ niedriger Energieverbrauch
A000063X
§ temperaturunabhängig
UN=Messspannung /
measuring voltage
§ 4 - 20 mA Ausgangssignal
§ fehlersicher
Stromquelle /
current source
§ keine Nullpunktdrift
Heizung / heater
§ hohe Messgenauigkeit
§ sehr lange Lebensdauer
Die Stromquelle wird eingeschaltet. Die Kammer wird evakuiert,
der Partialdruck P2 ändert sich linear mit der transportierten
Ladungsmenge. Dadurch steigt an der anderen Scheibe die
Spannung an. Wenn die Spannung einen bestimmten Wert
erreicht hat, ändert die Stromquelle ihre Stromrichtung, es
DD 55 0001
10/2011
Seite 1/3
®
§ flexibel einsetzbar
§ keine Referenzgase notwendig
§ keine Kalibriergase notwendig
Bühler Technologies GmbH
D - 40880 Ratingen, Harkortstr. 29
Tel.: + 49 (0) 2102 / 49 89-0 Fax: + 49 (0) 2102 / 49 89-20
Internet: www.buehler-technologies.com
e-mail: analyse@buehler-technologies.com
Technische Daten Sensor BA 1000 S
90
ca. 5 Minuten
0,1 - 25 Vol.% O2
4-20 mA
+/- 2 % vom Messbereichsendwert
von 0 °C bis + 450 °C
bis 10 m/s
12 mm
ca. 3 sec
- 10 bis + 50 °C
- 20 bis + 70 °C
24 V DC
650 mA
4,4 A
max.100 m
IP 54
85 x 90 x 65 mm
ca. 1,1 Kg
Aluminium
Edelstahl rostfrei
ø12
Aufheizzeit
Messbereich
Ausgang
Genauigkeit
Entnahmetemperatur
Abgasgeschwindigkeit
Sondenrohrdurchmesser
Reaktionszeit
Umgebungstemperatur
Lagertemperatur
Spannungsversorgung
Strom
Einschaltstrom
Entfernung Sonde / Elektronik
Schutzart
Abmessung ( B x H x T )
Gewicht
Werkstoff
Gehäuse
Sondenrohr
A000062X
A
71
L
65
Ansicht A
153
85
Ausgang
Eingang
Erdanschluss
Bestellhinweise:
Bezeichnung:
Sensor BA 1000 S Entnahmelänge L = 220 mm
Sensor BA 1000 S Entnahmelänge L = 380 mm
Sensor BA 1000 S Entnahmelänge L = 780 mm
Optionen
Netzteil BA 1000 N für Wandaufbau
Spannungsversorgung
Schutzart
Gewicht
Artikel-Nr.
55 01 398
55 01 399
55 01 499
55 01 599
115 / 230V AC 48 - 62 Hz wahlweise
IP 65
2,1 kg
Digitalanzeige BA 1000 D 230 G für Wandaufbau 4 ½ stellig, inkl. Netzteil
mit 2 Grenzwertkontakten Schaltfunktion min - max,
Schaltpunkt, Hysterese, Anzug- und Abfallzeiten frei einstellbar
Kontaktbelastbarkeit
250 V / 3 A
Spannungsversorgung
115 / 230V AC 48 - 62 Hz wahlweise
Schutzart
IP 65
Gewicht
2,7 kg
55 01 899
Kalibrieranzeige BA 1000 KA
zur schnellen Überprüfung der Kalibrierung vor Ort
Abmessung
Gewicht
55 01 2992
Zubehör :
Einschraubverschraubung
Einschraubverschraubung
Verbindungskabel 5 m
Vorfilter ohne Montageflansch
Vorfilter incl. Montageflansch
DD 55 0001
10/2011
Seite 2/3
54 x 100 x 40 mm (B x H x T)
0,12 kg
1/2“ NPT mit PTFE Klemmring
1/2“ NPT mit Edelstahl Klemmring
( andere Längen auf Anfrage )
55 01 2993
55 01 2994
55 01 2995
55 01 29971
55 01 2997
Technische Änderungen vorbehalten
DD 55 0001
10/2011
Seite 3/3
Zirkonia Oxygen Analyser
BA 1000
Description
The Buhler Model BA 1000 solid-state oxygen analyser has
been developed for in-situ applications. The cell is selfcontrolling and therefore well- suited for applications in flue
gas or similar matrices.
Principle of operation
The dynamic sensor measures oxygen without using a
reference gas. The sensor is consists of two zirconium discs
covering both sides of a platinum ring. An additional Al2O3
coating protects the sensor from environmental influence
and increases it´s life expectancy. The measurement
chamber (P2) is filled with a sample gas. One disc is
connected with a reversible source of current. It is then used
as an solid-state electronic oxygen pump. On the opposite
disc the partial pressure of oxygen is determined by
measuring the induced Nernst potential.
potential has reached a certain threshold, the polarity is
reversed. Oxygen ions are then pumped back into the
chamber and oxygen concentration increases. After
reaching a certain threshold, the polarity changes again.
This process is repeated over and over. The time for one
period is linearly proportional to the oxygen concentration in
the sample gas.
An optional pre-filter enables operation in gas streams with
high dust concentrations.
Fields of applications include:
Furnaces, industrial heating facilities, industrial combustion
facilities, farming (in silos), medical, biotechnology
Advantages and benefits
§ Low energy consumption
Sensor
A000063X
Glas / glass
§ Temperature independent
GND
UN=Messspannung /
measuring voltage
Stromquelle /
current source
Heizung / heater
First, the oxygen ions are pumped out of the chamber. The
change in electrical charge on the first disc changes in direct
linear proportion to the oxygen concentration. This increases
the electric potential at the opposite disc. When the electric
DE 55 0001
10/2011
Page 1/3
®
§
§
§
§
§
§
§
§
Linear output signals 4 - 20 mA
Error protected
High measurement precision
Independent of gas matrix
Long life
Variety of applications
No reference gas
No calibration gas
Bühler Technologies GmbH
D - 40880 Ratingen, Harkortstr. 29
Tel.: + 49 (0) 2102 / 49 89-0 Fax: + 49 (0) 2102 / 49 89-20
Internet: www.buehler-technologies.com
e-mail: analyse@buehler-technologies.com
Technical Data - Sensor BA 1000 S
90
approx. 5 minutes
0.1 - 25 Vol.% O2
4-20 mA
± 2 % of full scale value
0 °C up to +450 °C
up to 10 m/s
12 mm
approx. 3 seconds
-10 to +50 °C
-20 to +70 °C
24 V DC
650 mA
4.4 A
max.100 m
IP 54
85 x 90 x 65 mm
ca. 1.1 kg
stainless steel
aluminium painted
ø12
Warm-up time
Measurement range
Output
Accuracy
Sample temperature
Exhaust gas velocity
Probe diameter
Response time
Operating ambient temperature
Storage temperature
Power supply
Current
Switch-on current
Distance analyser/ control unit
Protection class
Dimensions ( W x H x D )
Weight
Material
Probe tube
Head
A000062X
A
71
L
65
View A
153
85
Output
Input
Protection ground
Please indicate with order:
Description:
Sensor BA 1000 S with probe length L = 220 mm
Sensor BA 1000 S with probe length L = 380 mm
Sensor BA 1000 S with probe length L = 780 mm
Options
Power supply BA 1000 N wall mounted
Power supply
Protection class
Weight
4 ½ digit LED display BA 1000 D 230 G wall mounted
With 2 alarms for min. and max. function
Free programmable
Relays
Power supply
Protection class
Weight
Calibration display BA 1000 KA
Quick-check calibration unit
Dimensions
Weight
Tools :
Pipe fitting for probe mounting
Pipe fitting for probe mounting
Interconnecting cable 5 m
Complete pre filter with mounting flange
Pre filter without mounting flange
DE 55 0001
10/2011
Page 2/3
Part-no.
55 01 398
55 01 399
55 01 499
55 01 599
115 / 230V AC 48 - 62 Hz alternatively
IP 65
2.1 kg
55 01 899
250 V / 3 A
115 / 230V AC 48 - 62 Hz alternatively
IP 65
2.7 kg
55 01 2992
54 x 100 x 40 mm (W x H x D)
0.12 kg
1/2“ NPT with PTFE locking ring
1/2“ NPT with stainless steel locking ring
( other lenght on request )
55 01 2993
55 01 2994
55 01 2995
55 01 2997
55 01 29971
we reserve the right to amend specification
DE 55 0001
10/2011
Page 3/3
Zirconia Oxygen Analyzer
BA 1000
Description
The Buhler Model BA 1000 solid-state oxygen analyzer has
been developed for in-situ applications. The cell is selfcontrolling and therefore well- suited for applications in flue
gas or similar matrices.
A000063X
Principle of operation
The dynamic sensor measures oxygen without using a
reference gas. The sensor is consists of two zirconium discs
covering both sides of a platinum ring. An additional Al2O3
coating protects the sensor from environmental influence
and increases it´s life expectancy. The measurement
chamber (P2) is filled with a sample gas. One disc is
connected with a reversible source of current. It is then used
as an solid-state electronic oxygen pump. On the opposite
disc the partial pressure of oxygen is determined by
measuring the induced Nernst potential.
First, the oxygen ions are pumped out of the chamber. The
change in electrical charge on the first disc changes in direct
linear proportion to the oxygen concentration. This increases
the electric potential at the opposite disc. When the electric
potential has reached a certain threshold, the polarity is
DA 55 0001
10/2011
Page 1/3
reversed. Oxygen ions are then pumped back into the
chamber and oxygen concentration increases. After
reaching a certain threshold, the polarity changes again. This
process is repeated over and over. The time for one period is
linearly proportional to the oxygen concentration in the
sample gas.
An optional pre-filter enables operation in gas streams with
high dust concentrations.
Fields of applications include:
Furnaces, industrial heating facilities, industrial combustion
facilities, farming (in silos), medical, biotechnology
Advantages and benefits
§
§
§
§
§
§
§
§
§
Low energy consumption
Temperature independent
Linear output signals 4 - 20 mA
Error protected
High measurement precision
Independent of gas matrix
Long life
Variety of applications
No reference gas
§ No calibration gas
Buhler Technologies LLC
1030 West Hamlin Road, Rochester Hills, MI 48309
Phone: 248.652.1546
Fax: 248.652.1598
Internet: www.buhlertech.com
e-mail: sales@buhlertech.com
3.54”
Warm-up time
approx. 5 minutes
Measurement range
0.1 - 25 Vol.% O2
Output
4-20 mA
Accuracy
+/- 2 % of full scale value
Sample temperature
32 °F up to + 840 °F
Exhaust gas velocity
up to 32 ft/s
Probe diameter
12 mm (0.47”)
Response time
approx. 3 seconds
Operating ambient temperature 15 to 120 °F
Storage temperature
- 5 to + 160 °F
Power supply
24 V DC
Current
650 mA
Switch on current
4.4 A
Distance analyzer/control unit max. 325 ft.
Protection class
IP 54
Dimensions ( B x H x T )
85 x 90 x 65 mm
3.3 x 3.5 x 2.6 in.
Weight
ca. 2.43 lb
Material
Probe tube
stainless steel
Head
aluminum painted
A000062X
ø0.47”
Technical Data - Sensor BA 1000 S
A
2.80”
L
2.56”
View A
6.02”
3.35”
Output
Input
Protection ground
Please indicate with order:
Description:
Sensor BA 1000 S with probe length L = 220 mm (8.7 in.)
Sensor BA 1000 S with probe length L = 380 mm (15.0 in.)
Sensor BA 1000 S with probe length L = 780 mm (30.7 in.)
Options
Power supply BA 1000 N wall mounted
Power supply
Protection class
Weight
4 ½ digit LED display BA 1000 D 230 G wall mounted
With 2 alarms for min. and max. function
Free programmable
Relays
Power supply
Protection class
Weight
Calibration display BA 1000 KA
Quick-check calibration unit
Dimensions
Weight
Tools :
Pipe fitting for probe mounting
Pipe fitting for probe mounting
Interconnecting cable 5 m
Complete pre filter with mounting flange
Pre filter without mounting flange
DA 55 0001
10/2011
Page 2/3
Part-no.
55 01 398
55 01 399
55 01 499
55 01 599
115 / 230V AC 48 - 62 Hz alternatively
IP 65
4.6 lb
55 01 899
250 V / 3 A
115 / 230V AC 48 - 62 Hz alternatively
IP 65
6.0 lb
55 01 2992
54 x 100 x 40 mm (W x H x D)
0.26 lb
1/2“ NPT with PTFE locking ring
1/2“ NPT with stainless steel locking ring
( other length on request )
55 01 2993
55 01 2994
55 01 2995
55 01 2997
55 01 29971
we reserve the right to amend specification
Note: All dimensions in this diagram are given in mm (unless otherwise noted)!
DA 55 0001
10/2011
Page 3/3
RMA - Dekontaminierungserklärung
RMA - Decontamination Statement
DE/EN
Gültig ab / valid since: 2014/11/01
Revision / Revision 1
ersetzt Rev. / replaces Rev. 0
Um eine schnelle und reibungslose Bearbeitung Ihres Anliegens zu erreichen, füllen Sie bitte diesen Rücksendeschein
aus. Eine genaue Fehlerbeschreibung ist für die Ursachenanalyse nötig und hilft bei der schnellen Bearbeitung des
Vorgangs. Die Aussage „Defekt“ hilft bei der Fehlersuche
leider nicht.
Please complete this return form to ensure your claim is
processed quickly and efficiently. An accurate description of
the problem is necessary for cause analysis and will help
processing the claim quickly. Unfortunately, stating
“defective” will not help us troubleshoot the issue.
Die RMA-Nummer bekommen Sie von Ihrem
Ansprechpartner im Vertrieb oder Service.
You may obtain the RMA number from your sales or
service representative.
Zu diesem Rücksendeschein gehört eine Dekontaminierungserklärung. Die gesetzlichen Vorschriften schreiben vor, dass
Sie uns diese Dekontaminierungserklärung ausgefüllt und
unterschrieben zurücksenden müssen. Bitte füllen Sie auch
diese im Sinne der Gesundheit unserer Mitarbeiter
vollständig aus.
This return form includes a decontamination statement. The
law requires you to submit this completed and signed
decontamination statement to us. Please complete the entire
form, also in the interest of our employees’ health.
Bringen Sie den Rücksendeschein mit der
Dekontaminierungserklärung bitte zusammen mit den
Versandpapieren in einer Klarsichthülle außen an der
Verpackung an. Ansonsten ist eine Bearbeitung Ihres
Reparaturauftrages nicht möglich!
Attach the return form including decontamination
statement along with the shipping documentation to the
outside of the package, inside a clear pouch. Otherwise
we are unable to process your repair order!
Angaben zum Absender:
Sender information:
Firma / Company
Ansprechpartner /
Contact person
Anschrift / Address
Abteilung /
Department
E-Mai / E-Mail:
Tel. / Phone
Fax / Fax:
Artikelnummer /
Item number
Auftragsnummer /
Order number
Anzahl / Quantity
Rücksendegrund /
Return reason
RMA-Nr. /
RMA no.
Reparatur / Repair
Garantie / Warranty
Zur Prüfung / For inspection
Rückgabe / Return
Vorgangsnummer des Kunden /
Customer transaction number::
Fehlerbeschreibung / Description of the problem:
Ort, Datum /
Place, Date
Seite
Page
1/2
Unterschrift / Stempel /
Signature / Stamp:
Bühler Technologies GmbH
D - 40880 Ratingen, Harkortstr. 29
Tel.: + 49 (0) 2102 / 4989-0 Fax: + 49 (0) 2102 / 4989-20
e-mail: service@buehler-technologies.com
Internet: www.buehler-technologies.com
RMA - Dekontaminierungserklärung
RMA - Decontamination Statement
DE/EN
Gültig ab / valid since: 2014/11/01
Revision / Revision 1
ersetzt Rev. / replaces Rev. 0
Bitte füllen Sie diese Dekontaminierungserklärung für jedes
einzelne Gerät aus.
Gerät / Device
RMA-Nr /
RMA no:
Serien-Nr. / Serial no.
[
]
Please complete this decontamination statement for each
individual item
Ich bestätige hiermit, dass das oben spezifizierte
Gerät ordnungsgemäß gereinigt und dekontaminiert
wurde und keinerlei Gefahren im Umgang mit dem
Produkt bestehen.
I herewith declare that the device as specified above has been
properly cleaned and decontaminated and that there are no
risks present when dealing with the device.
Ansonsten ist die mögliche Gefährdung genauer zu
beschreiben:
In other cases, please describe the hazards in detail:
Aggregatzustand (bitte ankreuzen):
Aggregate state (please check):
Flüssig / Liquid
Fest / Solid
Folgende Warnhinweise sind zu beachten
(bitte ankreuzen):
Pulvrig / Powdery
Please note the following warnings (please check):
Explosiv
Explosive
Giftig / Tödlich
Toxic / lethal
Entzündliche Stoffe
Flammable substances
Komprimierte Gase
Compressed gasses
Gesundheitsgefährdend
Hazardous to health
Gesundheitsschädlich
Harmful to health
Bitte legen Sie ein aktuelles Datenblatt des
Gefahrenstoffes bei!
Ort, Datum /
Place, Date:
Seite
Page
2/2
Gasförmig / Gaseous
Brandfördernd
Oxidizing
Umweltgefährdend
Harmful to the environment
Please include an updated data sheet of the hazardous
substance!
Unterschrift / Stempel
Signature / Stamp:
Bühler Technologies GmbH
D - 40880 Ratingen, Harkortstr. 29
Tel.: + 49 (0) 2102 / 4989-0 Fax: + 49 (0) 2102 / 4989-20
e-mail: service@buehler-technologies.com
Internet: www.buehler-technologies.com