Welker Automatic Insertion Probes IOM
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Installation, Operation, and
Maintenance Manual
Welker® Automatic Insertion Probe with
Outer Shaft Assembly & Horizontal Mount
Model
AIP-7L
The information in this manual has been carefully checked for accuracy and is intended to be used as a
guide for the installation, operation, and maintenance of the Welker equipment described above. Correct
operating and/or installation techniques, however, are the responsibility of the end user. Welker reserves
the right to make changes to this and all products in order to improve performance and reliability.
13839 West Bellfort
Sugar Land, TX 77498-1671
(281) 491-2331 - Office
(800) 776-7267 - USA Only
(281) 491-8344 - Fax
http://www.welkereng.com
© Copyright 2009, Welker, Inc.
TABLE OF CONTENTS
1.
1.1
1.2
1.3
1.4
2.
2.1
2.2
2.3
2.4
2.5
GENERAL
3
INTRODUCTION
DESCRIPTION OF PRODUCT
SPECIFICATIONS
SYSTEM COMPONENTS
3
3
4
5
GENERAL
PREPARING THE UNIT FOR INSTALLATION
INSTALLING THE UNIT
HELPFUL HINTS
RETRACTING THE UNIT
3.
MAINTENANCE
3.1
3.2
GENERAL
MAINTENANCE INSTRUCTIONS
4.1
TROUBLESHOOTING
6
7
9
10
11
13
13
13
APPENDIX A CROSS SECTIONAL DIAGRAM
4.
6
INSTALLATION & OPERATION
15
17
TROUBLESHOOTING GUIDE
17
Welker, Welker Jet®, and WelkerScope® are Registered Trademarks owned by Welker, Inc.
Document: IOM-099
AIP-7L
Rev: B
Last updated: 9/8/2009
Page 2 of 18
SPECIFICATIONS
1.
GENERAL
1.1
INTRODUCTION
We appreciate your business and your choice of Welker products. The installation,
operation, and maintenance liability for this product becomes that of the purchaser at the
time of receipt. Reading the applicable Installation, Operation, and Maintenance (IOM)
Manual prior to installation and operation of this equipment is required for a full
understanding of its application and performance prior to use.*
If you have any questions, please call 1-800-776-7267 in the USA or 1-281-491-2331.
Notes, Warnings, and Cautions
Notes emphasize information or set it off from the
surrounding text.
Caution messages appear before procedures that, if not observed, could
result in damage to equipment.
Warnings alert users to a specific procedure or practice
that, if not followed correctly, could cause personal injury.
*The following procedures have been written for use with standard Welker parts and equipment.
Assemblies that have been modified may have additional requirements and specifications that are not listed
in this manual.*
1.2 DESCRIPTION OF PRODUCT
The Welker Automatic Insertion Probes are designed for use in systems
where it is desirable to insert and retract a probe within a pipeline while the
pipeline remains pressurized. The design of the unit allows the operator to
control the movement of the probe into and out of the pipeline safely
through the use of five valves and an auxiliary gas supply, instrument air, or process fluid.
For gas sampling applications, the unit should be installed on top of a straight Section of piping
before the flowing stream is subjected to turns and impingements that can result in turbulent
flow. For liquid sampling applications, the probe should be installed in a Section of pipe where
the process fluid is properly conditioned, such as downstream of a mixing device.
Document: IOM-099
AIP-7L
Rev: B
Last updated: 9/8/2009
Page 3 of 18
SPECIFICATIONS
1.3
SPECIFICATIONS
The specifications listed in this Section are generalized for this equipment. Welker can modify the
equipment according to your company’s needs. However, please note that the specifications may
vary depending on the customization of your product.
Table 1
SPECIFICATIONS
Products
Materials of Construction
Insertion Length
Gases/Liquids
316 Stainless Steel, Carbon, Viton® and PTFE (others
available)
18", 24", 36" (45.72 cm, 60.96 cm, 91.44 cm)
(others available in 6" increments)
Viscosity Range
0.009 cp to 2,000 cp @ 68° F (20°C)
Pipeline Connection
2" ANSI Flange
(other connections available)
Sample Outlet Connection
1
Maximum Allowable Temperature *
325° F @ 500 psi
(163 ° C @ 35 bar)
Maximum Allowable Pressure *
1,440 psi @ -20° F to 100 ° F (Only at 600 ANSI)
(99 bar @ -29° C to 38° C)
/2" FNPT Need to Install a Valve
* Maximum allowable temperatures and pressures may be lower depending on specifications of pipeline connection device.
In a pipeline with liquid service, Welker strongly recommends using an auxiliary gas supply in order
to prevent damage to the cylinder and the oil reservoir.
Oil Reservoir
With the use of a hydraulic oil reservoir, process or auxiliary pressure is applied
within the reservoir in order to insert and retract it from the pipeline. The oil in
the reservoir is applied on the probe piston to assure a smooth travel. The
reservoir is shipped from the factory with the necessary oil volume. It should be
noted that the AIP-7L is also shipped from the factory with the assumption that
the installation will be horizontal. This unit is designed so the reservoir will be
placed in a vertical position, and the reservoir should remain in this position.
Document: IOM-099
AIP-7L
Rev: B
Last updated: 9/8/2009
Page 4 of 18
SPECIFICATIONS
1.4
SYSTEM COMPONENTS
•
•
•
•
•
•
•
•
Bottom Flange
Top Flange
Lock Collar
Probe Shaft (Outer)
Probe Shaft (Inner)
Outer Cap Assembly
Oil Fill Pot
Needle valves and associated piping, fittings, etc.
Document: IOM-099
AIP-7L
Rev: B
Last updated: 9/8/2009
Page 5 of 18
INSTALLATION
2.
INSTALLATION INSTRUCTIONS
2.1
GENERAL
After unpacking the unit, check it for compliance and for any damages that may have
occurred during shipment.
Claims for damages caused during shipping must be initiated by the receiver and directed to the
shipping carrier. Welker is not responsible for any damages caused from mishandling by the
shipping company.
When sealing fittings with PTFE tape, refer to the proper sealing instructions for the tape used.
The Welker AIP-7L was designed according to customer specifications. This manual contains
photographs of a standard Welker Automatic Insertion Probe, and are provided for part
identification guidance only. When necessary, please refer to assembly drawing in Appendix A
for specific part placement and design.
Recommended Tools
It would be advisable to have the following tools available
for installation of the unit; however, tools used will vary
depending on probe model.
• Measuring tape
• Small hex key set
• 6" adjustable wrench
• 10" adjustable wrench
• Flat blade screwdriver
• Permanent marker
Document: IOM-099
AIP-7L
Rev: B
Last updated: 9/8/2009
Page 6 of 18
INSTALLATION
2.2
1.
PREPARING THE UNIT FOR INSTALLATION
Determine the insertion length
Before installing the probe, the length the insertion probe will need to travel inside the
pipeline must be determined. Measure from the top of the pipeline’s isolation valve to the
center 1/3 of the pipeline (Insertion Length A on Figure 2)
.
Not retracted
Figure 3- Probe Retraction
Figure 2 - Insertion Length
A
2.
Retracted
Set the insertion length on the probe
Once the insertion length of the probe is determined, this length should be measured on
the probe itself.
• Pull up on the probe to make sure it is fully retracted (see Figure 3).
• Begin at the top cap, and measure up on the probe to the desired length (see Figure 4).
• Mark this point on the probe, as this is where the bottom of the lock collar will be
positioned.
Valve "D"
Figure 4 Insertion
Length
Valve "A"
Valve "D"
Valve "C"
Figure 5 - Horizontal Probe Mount
Document: IOM-099
AIP-7L
Rev: B
Last updated: 9/8/2009
Valve "B"
Page 7 of 18
INSTALLATION
2.2
PREPARING THE UNIT FOR INSTALLATION (CONTINUED)
3.
Position the lock collar
• Loosen the two set screws in the lock collar, and move the bottom of the collar to the
position noted in the previous step.
• Position the lock screw so that it is aligned with its port in the top cap.
• Tighten the lock collar set screws.
4.
If necessary, connect an instrument air or an auxiliary gas supply
If the pipeline contains a liquid or toxic gas, you will need to use an alternative source for
pressure.
• Detach the tubing between Valve B and Valve A.
• Remove Valve A from the base cap or flange of the probe.
• Plug the opening that is now in the base cap or flange with a 1/4" NPT plug.
• Connect the auxiliary gas supply to Valve B.
5.
Install a 1/2" NPT valve on the outlet cap assembly.
Document: IOM-099
AIP-7L
Rev: B
Last updated: 9/8/2009
Page 8 of 18
INSTALLATION
2.3
1.
2.
3.
4.
INSTALLING THE UNIT
Close all valves on the AIP-7L including the 1/2" outlet valve.
Connect the unit to the pipeline via the pipeline isolation valve.
Slowly open the pipeline isolation valve, and check for leaks. If you are using an auxiliary
gas supply, open the supply valve.
Open Valves A and B on the unit to allow pipeline pressure to enter the oil reservoir.
Valve D should always be closed when pipeline pressure is flowing through Valves A and B.
Opening Valve D while pipeline pressure is flowing may cause the oil in the reservoir to erupt
from the valve.
4.
Slowly open Valves C & E, located between the oil reservoir and the top cap. The probe
will begin to insert into the pipeline (see Figure 6).
Opening the valve too quickly may cause the probe to insert abruptly into the pipeline, possibly
resulting in damage to the unit.
Once the probe begins to move, there is no need to open
the valve any further. This assures a slow and smooth
insertion of the probe into the pipeline.
5.
6.
7.
8.
9.
When the lock collar reaches the top cap, close valve C.
Tighten the lock collar and hold down the nuts to the top flange.
Close all valves.
Check the entire system for leaks. The unit is now in service.
You may now attach the 1/2" outlet valve.
Document: IOM-099
AIP-7L
Rev: B
Last updated: 9/8/2009
Page 9 of 18
INSTALLATION
2.4
HELPFUL HINTS
1.
Avoid rough handling of the unit and bending of the probe. The probe has a polished
surface that travels through seals.
2.
Operate the unit slowly and smoothly while inserting and retracting to avoid unnecessary
slamming of the lock collar and/or the probe piston located inside the unit.
3.
The most common cause for repairs to an automatic insertion probe is due to the pipeline
isolation valve closing on the probe while the probe is still inserted into the pipeline.
Please avoid this practice.
4.
The entire unit should be treated with care.
Document: IOM-099
AIP-7L
Rev: B
Last updated: 9/8/2009
Page 10 of 18
INSTALLATION
2.5
RETRACTING THE UNIT
The instrument attached to the unit must be shut down or disconnected prior to retracting.
Make sure all valves on the unit are closed prior to installation or removal.
Recommended Tools
It would be advisable to have the following tools available for retracting the unit;
however, tools used will vary depending on probe model.
•
•
•
•
1.
2.
3.
4.
Small hex key set
6" adjustable wrench
10" adjustable wrench
Flat blade screwdriver
With all valves closed, open valves A, E, and C. This will assure the probe is ready for
step 3, removing the lock collar.
Remove any tubing attached to the 1/2" outlet valve.
Close all valves, including A, E, and C.
Loosen the lock collar screw and hold down the nuts. At this point, the probe will remain
inside the pipeline.
Document: IOM-099
AIP-7L
Rev: B
Last updated: 9/8/2009
Page 11 of 18
INSTALLATION
2.5
RETRACTING THE UNIT (CONTINUED)
Valves A and B should always be closed when opening Valve D. Opening Valve D while pipeline
pressure is flowing may cause the oil in the reservoir to erupt from the valve.
5.
Slowly open Valve D. This will vent the air inside the oil reservoir. Leave this valve open.
6.
Slowly open Valve C to allow pressure to be relieved from the insertion probe. The probe
will now begin to retract from the pipeline.
Once the probe begins to move, Valve C does not need to be opened any further. This assures a slow
and smooth retraction of the probe.
7.
When the probe has completely retracted from the pipeline, close the pipeline isolation
valve, cutting off any pressure flowing into the unit.
8.
If you are using instrument air or an auxiliary gas supply, close the supply valve.
If the probe needs to be withdrawn from the pipeline but the unit itself does not need to be removed
from the pipeline, stop at step 8 of this Section.
9.
Vent the pressure in the unit by opening Valves E, A, and B allowing pressure to be
released, and then close all of the valves.
If pressure does not stop venting from one or more of the unit’s valves, the pipeline isolation valve is
possibly leaking. User should consult on-sight safety engineer.
10.
Remove the unit from the pipeline isolation valve.
11.
The unit is now ready for maintenance or to be moved to another location.
Document: IOM-099
AIP-7L
Rev: B
Last updated: 9/8/2009
Page 12 of 18
MAINTENANCE
3.
MAINTENANCE
3.1
GENERAL
Prior to maintenance or disassembly of the unit, it is advisable to have a repair kit handy for
the system in case of unexpected wear or faulty seals. All maintenance and cleaning of the
unit should be done on a smooth, clean surface.
We recommend that the unit have annual maintenance under normal operating conditions. In the case
of severe service, dirty conditions, excessive cycling usage, or other unique applications that may
subject the equipment to unpredictable circumstances, a more frequent maintenance schedule may be
appropriate.
3.2
MAINTENANCE INSTRUCTIONS
Refer to Appendix A throughout the entire maintenance process.
Recommended Tools
It would be advisable to have the following tools available for maintenance of the unit;
however, tools used will vary depending on probe model.
•
•
•
•
•
•
Small hex key set
6" adjustable wrench
10" adjustable wrench
10" channel lock pliers
Fine-grit sandpaper
Flat blade screwdriver
1.
Close all valves.
2.
Disconnect and remove all tubing.
3.
Remove the oil reservoir and mounting clamp.
4.
Unclamp the top lock collar clamp (Part 29).
The locking pin (part 28) is very small. Be careful not to drop it, and keep it in a safe place for
reassembly.
5.
Remove the eight nuts and bolts and the lock collar (Part 12) and top flange (Part 14).
6.
Remove the inlet cap (Part 10) and the outlet cap (Part 1).
Document: IOM-099
AIP-7L
Rev: B
Last updated: 9/8/2009
Page 13 of 18
MAINTENANCE
3.2
MAINTENANCE INSTRUCTIONS (CONTINUED)
Refer to Appendix A throughout the entire maintenance process.
7.
Slide out the outer shaft assembly (part 11) from the cylinder.
8.
Remove and replace the seals and O-rings (Parts 7, 17, 15, 5, and 2) if necessary.
9.
Replacing the seals
Do not dig into the metal surfaces of the parts when removing O-rings from the O-ring grooves. Scratching
the sealing surface can result in a leak. If necessary, dig into the O-ring, and replace it during reassembly. If
the sealing surface becomes damaged, use a 600-grit wet sand paper strip to smooth the surface, and then
clean it.
New seals supplied in spare parts kits are not lubricated. They should be lightly coated with lubrication
grease (Dow Corning 111 [DC 111] grease or equivalent lubricant) before they are installed into the
equipment. This helps with the installation of the seals while reducing the risk of damage when positioning
them on the parts. After the seals are installed, some additional lubrication can be applied to the probe or
cylinder inner diameters to allow smooth transition of parts.
10.
Examine the outer and inner surfaces of the cylinder and inner shaft for a smooth finish. If
there are any pits or major scratches, the seals will leak. Call Welker for service if necessary.
11.
Remove and replace the seals in the inlet cap (Parts 19 and 20).
12.
Carefully, slide the inner shaft back into the cylinder, replace the cylinder, and replace the
inlet cap.
13.
Replace the top flange, lock collar, and eight nuts and bolts.
14.
Replace the top lock collar.
15.
Re-clamp the oil reservoir and replace all tubing.
16. Add oil to the oil reservoir
The unit is shipped from the factory with the necessary oil volume. If oil is needed, remove
C until the reservoir is ¾ full. Replace Valve D.
Valve D and add oil
D
If you are adding oil while the unit is still assembled and attached to a pipeline, depressurize the assembly,
making sure Valve B is closed before removing Valve D.
B
If oil needs to be added, it may be due to a leak in the unit.
Document: IOM-099
AIP-7L
Rev: B
Last updated: 9/8/2009
Page 14 of 18
APPENDIX A CROSS SECTIONAL
DIAGRAM
Document: IOM-099
AIP-7L
Rev: B
Last updated: 9/8/2009
Page 15 of 18
APPENDIX A CROSS SECTIONAL
DIAGRAM
Document: IOM-099
AIP-7L
Rev: B
Last updated: 9/8/2009
Page 16 of 18
TROUBLESHOOTING
4.1 TROUBLESHOOTING GUIDE
The following is a troubleshooting table of issues most commonly associated with the
Welker Liquid Knockout System models. If you are having a problem that is not listed,
or if the solution provided does not repair the problem, please call Welker for service
options.
POSSIBLE
CAUSE
PROBLEM
SOLUTION
The probe doesn’t insert or
retract smoothly.
Air may be trapped in the oil
reservoir, or the reservoir may
need oil.
Check to make sure the oil
reservoir is ¾ full; vent any air
trapped in the reservoir.
See step four in Section 3.3.
See step four in Section 2.5.
The oil reservoir needs to be
refilled often.
Oil may be leaking past the
piston seal.
Replace the piston seal, and
reassemble the probe.
See step 3 in Section 3.3.
See Section 3.2 and 3.4.
Pipeline or process pressure is
leaking from the base cap’s
vent hole.
Seals in the base cap are
leaking.
Replace the seals.
See step 3 in Section 3.3.
See Figure 14.
•
• Gently pull up on the probe
The probe will not retract
from the pipeline.
There may not be enough
pressure in the pipeline to
eject the probe.
• The probe is bent inside the
pipeline, possibly due to
pipeline velocity or the
isolation valve closing on
the probe while the probe
is still inserted in the
pipeline.
Document: IOM-099
AIP-7L
Rev: B
Last updated: 9/8/2009
until it begins to retract.
•
The unit will need to be
repaired or replaced. Call
Welker for service options.
Page 17 of 18
13839 West Bellfort, Sugar Land, Texas 77498-1671
Phone: (281) 491-2331
Fax: (281) 491-8344
Toll Free: (800) 776-7267
Web Page: www.welkereng.com