Yokogawa User's Manual Model VD6SM Sampling Unit For Vibration Type Density Meter
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
User’s
Manual
Model VD6SM
Sampling Unit
For Vibration Type Density Meter
IM 12T3S1-01E
R
IM 12T3S1-01E
3rd Edition
Blank Page
i
INTRODUCTION
The Model VD6SM sampling unit for the vibration type density meter is used for:
•
Controlling the properties (e.g., the flow rate and pressure) of measurement liquid to meet the
specifications of the Model VD6D (or VD6DF) density detector.
•
Ensuring that the properties of measurement liquid is always in good condition and that normal
operation is maintained.
Contents of This Manual and Related User’s Manual
This user’s manual provides instructions on how to use the VD6SM sampling unit for the vibration type
density meter and only related information in regard to the Model DM8 vibration type liquid density
meter.
Note: The Model DM8 vibration type liquid density meter consists of a density detector (e.g., Model
VD6D or VD6DF), a density converter (Model DM8C) and a dedicated cable (Model DM8W).
To use the Model VD6SM sampling unit for the vibration type density meter properly, read this manual
thoroughly.
As for the Model DM8 vibration type liquid density meter, refer to User’s Manual, IM 12T03A01-02E.
INSTRUCTIONS FOR USE
CAUTION
Restriction on application
The Model VD6SM sampling unit for the vibration type density meter is not suited to be used for
online measurement in the food industry due to its structure.
Use of a flameproof detector
When the meter is installed in an area where combustible gases may exist, the Model VD6DF
flameproof detector should be used. For its use, restriction as a flameproof instrument should be
strictly observed. For details of restrictions, refer to the user’s manual of the Model DM8 vibration
type liquid density meter, IM 12T03A01-02E.
High temperature and pressure measurement liquid
Maximum temperature and pressure of the measurement liquid that can be supplied to the
VD6SM sampling unit are 100 °C and 2 MPa, respectively. When high temperature and pressure
liquid is supplied to the sampling unit, it is recommended that use of such liquid should be clearly
indicated to draw maintenance engineers’ attention. Never supply high temperature and pressure liquid that exceeds the limitations.
Steam trace
When a steam trace piping is required, install it in accordance with Yokogawa engineer’s advice
and the information described in this manual. Improper steam trace may cause measurement
errors.
Prohibition of modification
Repair of the measurement liquid piping is subject to restoration to the original form.
For part replacement, be sure to use the specified parts.
Media No. IM 12T3S1-01E
3rd Edition : Jul. 2012 (YK)
All Rights Reserved Copyright © 1988, Yokogawa Electric Corporation
IM 12T3S1-01E
ii
UNPACKING AND SPECIFICATION CHECK
The VD6SM sampling unit includes the customer specified density detector (ordered separately)
when shipped from the factory. The VD6SM weighs approximately 80 kg (varies depending on the
specifications). It is recommended that the unit be unpacked near the installation location.
CAUTION
Instructions for Unpacking and Transportation
The unit should be unpacked on a flat place while keeping it in a position of installation. (Avoid
strong impact, e.g., overturn).
Do not apply force to the piping, otherwise connection failure may result.
After unpacking, ensure that the product received meets specifications. The specifications can be
checked by the Model and Suffix Code indicated on the name plate at the front of the stand. For details of the Model and Suffix Code, refer to Section 1.3.
Note: The specification of the density detector should be checked by the Model and Suffix Code
indicated on the name plate on the terminal box. For details of the Model and Suffix Code, refer
to User’s Manual, IM 12T03A01-02E.
IM 12T3S1-01E
iii
For the safe use of this equipment
Safety, Protection, and Modification of the Product
•
In order to protect the system controlled by the product and the product itself and ensure safe
operation, observe the safety precautions described in this user’s manual. We assume no liability
for safety if users fail to observe these instructions when operating the product.
•
If this instrument is used in a manner not specified in this user’s manual, the protection provided
by this instrument may be impaired.
•
Be sure to use the spare parts approved by Yokogawa Electric Corporation (hereafter simply
referred to as YOKOGAWA) when replacing parts or consumables.
•
Modification of the product is strictly prohibited.
•
The following symbols are used in the product and user’s manual to indicate that there are
precautions for safety:
Notes on Handling User’s Manuals
•
Please hand over the user’s manuals to your end users so that they can keep the user’s manuals
on hand for convenient reference.
•
Please read the information thoroughly before using the product.
•
The purpose of these user’s manuals is not to warrant that the product is well suited to any
particular purpose but rather to describe the functional details of the product.
•
No part of the user’s manuals may be transferred or reproduced without prior written consent
from YOKOGAWA.
•
YOKOGAWA reserves the right to make improvements in the user’s manuals and product at any
time, without notice or obligation.
•
If you have any questions, or you find mistakes or omissions in the user’s manuals, please
contact our sales representative or your local distributor.
Warning and Disclaimer
The product is provided on an “as is” basis. YOKOGAWA shall have neither liability nor responsibility
to any person or entity with respect to any direct or indirect loss or damage arising from using the
product or any defect of the product that YOKOGAWA can not predict in advance.
IM 12T3S1-01E
iv
Symbol Marks
Throughout this user’s manual, you will find several different types of symbols are used to identify different sections of text. This section describes these icons.
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate
injury. It may also be used to alert against unsafe practices.
IMPORTANT
Indicates that operating the hardware or software in this manner may damage it or lead to system
failure.
NOTE
Draws attention to information essential for understanding the operation and features.
Tip
This symbol gives information that complements the current topic.
SEE ALSO
This symbol identifies a source to be referred to.
IM 12T3S1-01E
v
After-sales Warranty
Do not modify the product.
During the warranty period, for repair under warranty consult the local sales
representative or service office. Yokogawa will replace or repair any damaged
parts. Before consulting for repair under warranty, provide us with the model
name and serial number and a description of the problem. Any diagrams or
data explaining the problem would also be appreciated.
If we replace the product with a new one, we won’t provide you with a repair report.
Yokogawa warrants the product for the period stated in the pre-purchase quotation Yokogawa
shall conduct defined warranty service based on its standard. When the customer site is located
outside of the service area, a fee for dispatching the maintenance engineer will be charged to the
customer.
In the following cases, customer will be charged repair fee regardless of war-
ranty period.
•
Failure of components which are out of scope of warranty stated in instruction manual.
•
Failure caused by usage of software, hardware or auxiliary equipment, which Yokogawa Electric
did not supply.
•
Failure due to improper or insufficient maintenance by user.
•
Failure due to modification, misuse or outside-of-specifications operation which Yokogawa does
not authorize.
•
Failure due to power supply (voltage, frequency) being outside specifications or abnormal.
•
Failure caused by any usage out of scope of recommended usage.
•
Any damage from fire, earthquake, storms and floods, lightning, disturbances, riots, warfare,
radiation and other natural changes.
Yokogawa does not warrant conformance with the specific application at the
user site. Yokogawa will not bear direct/indirect responsibility for damage due
to a specific application.
Yokogawa Electric will not bear responsibility when the user configures the
product into systems or resells the product.
Maintenance service and supplying repair parts will be covered for five years
after the production ends. For repair for this product, please contact the nearest sales office described in this instruction manual.
IM 12T3S1-01E
Blank Page
v
Model VD6SM
Sampling Unit
For Vibration Type Density Meter
IM 12T3S1-01E 3rd Edition
CONTENTS
INTRODUCTION ...................................................................................................i
For the safe use of this equipment ..................................................................iii
After-sales Warranty ...........................................................................................v
1.
2.
GENERAL .................................................................................................. 1-1
1.1
Standard Specifications ................................................................................... 1-1
1.2
Accessories ....................................................................................................... 1-1
1.3
Models and Suffix Codes ................................................................................. 1-2
1.4
External Dimensions ........................................................................................ 1-3
INSTALLATION, PIPING AND WIRING.................................................... 2-1
2.1
2.2
2.3
3.
Installation ......................................................................................................... 2-1
2.1.1
Installation Area.................................................................................. 2-1
2.1.2
Installation .......................................................................................... 2-2
Piping ................................................................................................................. 2-3
2.2.1
Sampling line...................................................................................... 2-3
2.2.2
Steam Piping...................................................................................... 2-4
2.2.3
Cleaning Air Pipeline .......................................................................... 2-5
Wiring ................................................................................................................. 2-6
2.3.1
Special Cable ..................................................................................... 2-6
2.3.2
Wiiring Instructions............................................................................. 2-6
OPERATION .............................................................................................. 3-1
3.1
Names and Functions of Components........................................................... 3-1
3.2
Preparations for Operation .............................................................................. 3-2
3.3
3.4
3.2.1
Removing Shipping Parts from Density Detector .............................. 3-2
3.2.2
Inspecting Piping and Wiring Connections ........................................ 3-2
Start-Up .............................................................................................................. 3-3
3.3.1
Connecting Power ............................................................................. 3-3
3.3.2
Sampling Measurement Liquid .......................................................... 3-3
3.3.3
Calibration .......................................................................................... 3-4
3.3.4
Measuring the Pressure Compensation Valves ................................ 3-5
3.3.5
Operation Check for Complete Density Measurement System ........ 3-5
Normal Operation.............................................................................................. 3-6
3.4.1
Instructions for Operation................................................................... 3-6
3.4.2
Cleaning Strainer Element ................................................................. 3-6
IM 12T3S1-01E
vi
3.4.3
Checking Measured Density Value.................................................... 3-7
3.4.4
Cleaning Vibrator in the Density Detector.......................................... 3-8
3.4.5
Cautions When Stopping the Operation ............................................ 3-8
3.4.6
Consumable Part Replacement ........................................................ 3-8
Customer Maintenance Parts List ...................................... CMPL 12T03S01-02E
Revision Information ...............................................................................................i
IM 12T3S1-01E
1-1
< 1. GENERAL >
1.
GENERAL
The Model VD6SM sampling unit for the vibration type density meter controls the flow rate and
pressure of the liquid to be measured to meet the specifications of the Model VD6D (or VD6DF)
density detector, and also provides an alarm if conditions are abnormal.
1.1
Standard Specifications
External Dimensions:
Approx. 400 (W) x 400 (D) x 1350 (H) mm
Coating Finish:
Epoxy resin, baked gray finish (equivalent to Munsell N7)
Wetted Part Material:
Stainless steel JIS SUS316, Teflon TM
(gasket for flowmeter, pressure gauge and strainer), Ni for /FN option.
Process Conditions:
Inlet Temperature:
0 to 100°C
Inlet Pressure: 0 to 1 MPa or 0 to 2 MPa
Required Differential Pressure:
Flow Rate:
At least 0.1 MPa
1 to 10 l/min
Weight:
Approx. 80 kg
Process connection:
Screw, flange, or welding socket
Density Detector Used with the Sampling Unit:
Model VD6D and Model VD6DF.
(Note) This sampling unit cannot be used with the Model VD6DS density detector.
Ambient Temperature: -10 to 50°C
(Note) Depends on the density detector specifications.
Steam Tracing:
Available
Temperature:
1.2
I40 to 160°C
Accessories
Option code
without /FN
with /FN
Part Name
Part No.
Quantity
Strainer element
K9727KX
1
SUS316
Description
Gasket
K9727KW
1
Teflon
Strainer element
K9727LC
1
Ni
Gasket
K9727KW
1
Teflon
IM 12T3S1-01E
1-2
< 1. GENERAL >
1.3
Models and Suffix Codes
Model
Suffix Code
Description
VD6SM
••••••••••••••••••••••••••••••••••
•••••••••••••••••• Sampling unit for Vibration type Density
Meter (Note 1)
Piping connection
-JPT
-10K
-20K
-150
-300
-151
-301
-WST
••••••••••••••••••
••••••••••••••••••
••••••••••••••••••
••••••••••••••••••
••••••••••••••••••
••••••••••••••••••
••••••••••••••••••
••••••••••••••••••
Rc1/2
JIS 10K 15 RF Flange
JIS 20K 15 RF Flange
ANSI Class 150 1/2 RF Flange
ANSI Class 300 1/2 RF Flange
JPI Class 150 1/2 RF Flange
JPI Class 300 1/2 RF Flange
1/2 inch welding socket
••••••••••••••••••
••••••••••••••••••
••••••••••••••••••
••••••••••••••••••
1 MPa
2 MPa
Diaphragm type 1 MPa
Diaphragm type 2 MPa
Pressure guage
range
-PG10
-PG20
-PK10
-PK20
Temperature
range
Style code
Option
Material of strainer element
(Note 1)
(Note 2)
(Note 3)
IM 12T3S1-01E
Option Code
•••••••••••••••••• 0 to 100°C
•••••••••••••••••• 0 to 150°C
-T100
-T150
*B
•••••••••••••••••• Style B
/ST
/FN
With steam tracing (Note 2)
Ni (Note 3)
VD6SM Sampling unit is not including Detector. Order detector VD6D or VD6DF, separately.
DM8C converter and special cable DM8W are also required for sampling system of density meter.
If stean tracing is necessary, select the diaphragm type pressure guage.
If measuring solution includes NaOH (≤30%) , select option code /FN of Ni.
1-3
< 1. GENERAL >
1.4
External Dimensions
Unit : mm
NV3
NV4
NV2
270 400
4 - ø15
holes
For Anchor
Bolts
50
50
300
400
Pressure
Gauge
Thermometer
Detector
1350
Neeale
Flow Meter
Sample
Outlet
Sample
Inlet
NV1
strainer
200
34
Approx. 407
L
L
100
200
(100)
(for flange) (for welding
socket)
Model and Codes
VD6SM - JPT - PG
0 - T1
0 *B
VD6SM - 10K - PG
0 - T1
0 *B
VD6SM - 20K - PG
0 - T1
0 *B
VD6SM - 150 - PG
0 - T1
0 *B
VD6SM - 300 - PG
0 - T1
0 *B
VD6SM - 151 - PG
0 - T1
0 *B
VD6SM - 301 - PG
0 - T1
0 *B
VD6SM - WST - PG
0 - T1
0 *B
Thermometer
T
Connection Type
Rc 1/2 female
JIS 10K 15 RF Flange
JIS 20K 15 RF Flange
ANSI Class 150 1/2 RF Flange
ANSI Class 300 1/2 RF Flange
JPI Class 150 1/2 RF Flange
JPI Class 300 1/2 RF Flange
1/2 B Welding Socket
L
0
100
100
100
100
100
100
100
Pressure Gauge
Flow Meter
NV4
P
Needle Valve
Ball Valve
Sample
Inlet
strainer
Needle Valve
Needle Valve
NV2
Needle Valve
Detector
NV3
NV1
Sample
Outlet
Fig. 1.1 VD6SM - -PG 0- T1 0 *B Sampling Unit
IM 12T3S1-01E
1-4
< 1. GENERAL >
Unit : mm
NV3
NV4
NV2
4 - ø15
holes
50
P - P view
(Steam trace only)
400
For Anchor
Bolts
Steam Trap
V2
50
300
V1
270 400
Note: Some detail of steam trace
tube omitted in this drawings.
Pressure
Gauge
Thermometer
Detector
1350
Condensate
Outlet
Flow Meter
P
P
125 150
Steam
Inlet
Sample
Outlet
Sample
Inlet
NV1
150
strainer
200
34
Approx.407
L
100
L
(100)
200
(for flange) (for welding
socket)
Model and Codes
VD6SM - JPT - PK
0 - T1
0 *B /ST
VD6SM - 10K - PK
0 - T1
0 *B /ST
VD6SM - 20K - PK
0 - T1
0 *B /ST
VD6SM - 150 - PK
0 - T1
0 *B /ST
VD6SM - 300 - PK
0 - T1
0 *B /ST
VD6SM - 151 - PK
0 - T1
0 *B /ST
VD6SM - 301 - PK
0 - T1
0 *B /ST
VD6SM - WST - PK
0 - T1
0 *B /ST
Thermometer
Pressure Gauge
Flow Meter
T
Connection Type
Rc 1/2 female
JIS 10K 15 RF Flange
JIS 20K 15 RF Flange
ANSI Class 150 1/2 RF Flange
ANSI Class 300 1/2 RF Flange
JPI Class 150 1/2 RF Flange
JPI Class 300 1/2 RF Flange
1/2 B Welding Socket
b
P
b'
Sample
Inlet
a
Strainer
Needle Valve
NV2
Needle Valve
Needle Valve
Detector
NV1
Sample a'
Outlet
b
a
V1
Stop
Valve
V2
Steam
Inlet
b'
Condensate
Outlet
a'
Steam Trap
Fig. 1.2 VD6SM - -PK 0- T1 0 *B / ST Sampling Unit
IM 12T3S1-01E
NV4
Needle Valve
Ball Valve
NV3
L
0
100
100
100
100
100
100
100
2-1
< 2. INSTALLATION, PIPING AND WIRING >
2.
INSTALLATION, PIPING AND WIRING
2.1
Installation
2.1.1
Installation Area
The model VD6SM sampling unit for vibration type density meter should be installed in an area:
(1)
Near to the sampling point.
(2)
Free from mechanical vibration.
(3)
Free from corrosive gases and dust.
(4)
Free from excessive humidity.
(5)
Not subject to large temperature variations.
(6)
If the unit is used in a hazardous area, the prescribed operating ambient conditions must be
satisfied.
In such a case, the Model VD6DF Density detector is used.
WARNING
Restrictions When Installed in a Hazardous Area
A sampling unit including the VD6DF flameproof density detector can only be installed in a
hazardous area. The flameproof specification of the VD6DF density detector is TIIS d2G3 or
NEC Class I, Division 1, Groups C and D. Install the detector in an environment that meets these
flameproof specifications.
IM 12T3S1-01E
2-2
2.1.2
< 2. INSTALLATION, PIPING AND WIRING >
Installation
Normally, the sampling unit for the vibration type density meter is securely fixed to a well drained
concrete base (see Figure 2.1). When the unit might be exposed to the rain or direct sunlight, it should
be installed in a shelter.
Density meter sampling unit
Anchor bolt
Drain pit
Concrete base
Figure 2.1
IM 12T3S1-01E
Installation Example
< 2. INSTALLATION, PIPING AND WIRING >
2.2
2-3
Piping
The sampling unit for vibration type density meter requires the following piping:
(1) Sampling line
(2) Steam piping
However, the steam piping is necessary only when steam tracing is carried out. If necessary, perform
air piping as described in Sec. 2.2.3.
2.2.1
Sampling line
The sampling line transports the liquid to measure to the sampling unit. The main process pipe is
provided with a sample inlet and outlet. The sample conduit from the sample inlet is connected to
“SAMPLE IN” of the sampling unit, and the sample conduit from the outlet to “SAMPLE OUT” of the
sampling unit. The piping procedure is as follows:
(1)
The sample inlet and the sample outlet should be selected such that the pressure difference of
the liquid measured is at least 0.1 MPa.
(2)
The pipe length at the sample inlet side should not exceed 10 m.
(3)
Install tap and drain valves near the sampling unit (See Figure 2.2).
Main process pipe line
Pin (Pressure at inlet)
Pout (Pressure at outlet)
2 MPa>Pin Pout + 0.1 MPa
Sample conduits
(10 m max.)
Tap valve
Drain valve
Stop valve
Drain valve
Sampling unit
for vibration type density meter
Figure 2.2
(4)
Sample Line Connections
Avoid bending the pipe, or sludge may accumulate (See Figure 2.3).
Sampling unit
Main pipe
Sampling line
Sludge is likely to collect here.
Figure 2.3
Example of Unsatisfactory Piping
IM 12T3S1-01E
2-4
2.2.2
< 2. INSTALLATION, PIPING AND WIRING >
Steam Piping
This piping is required when the viscosity or fluid point of the liquid measured is high and the fluidity
must be increased. In this case, a tracer pipe is attached to the sampling unit.
The steam piping connects the steam heating source and the sampling unit “STEAM IN” inlet. The
sample conduit should be provided with a tracer pipe to which steam should be supplied. A steam
source of pressure 0.3 to 0.5 MPa and temperature 140 to 160 ºC is recommended.
The tracer pipe should be laid along the sampling line. Carried out after pressure leakage tests for the
sampling line have been completed.
For piping observe the following items:
(1)
The tracer pipe for the sample conducts should be installed such that all the piping is heated.
The pipe should also be wrapped with insulating material (See Figure 2.4).
Heat insulating material
Trace pipe
Sampling line
Figure 2.4
(2)
Steam Trace Section
The end of the “CONDENSASTE OUT” pipeline of the sampling unit should be open to the
atmosphere (See Figure 2.5).
Steam trap
Sampling unit
Drain pit
Figure 2.5
IM 12T3S1-01E
Piping at Steam Trap Discharge Side
< 2. INSTALLATION, PIPING AND WIRING >
2.2.3
2-5
Cleaning Air Pipeline
When a liquid which may stain the vibrator – stain on the vibrator results in measurement error – is
being measured in the density detector, periodical vibrator cleaning is required. When cleaning the
detector or when calibrating the density detector using standard solution instead of the liquid to be
measured, air is used to discharge the liquid in the vibrator. If no air supply is available in the vicinity of
the sampling unit installation area, it is recommended that you pipe air to the installation area.
The cleaning air pipe should be connected to the sampling unit only when vibrator cleaning or when
calibration using standard solution is being carried out. For the pipe for the air supply (clean dry air
with pressure 0.3 to 0.5 MPa G), a stop valve and a reducing valve should be provided near the
sampling unit, and a flexible, tube should be used to connect to the steel pipe (outside diameter 10
mm).
Sampling unit
BV1
DD
NV2
NV4
NV3
Connect when cleaning the density detector
Air supply 0.3 to 0.5 MPa G
Flexible tube
Figure 2.6
Reducing Stop valve
valve
Cleaning Air Piping
IM 12T3S1-01E
2-6
< 2. INSTALLATION, PIPING AND WIRING >
2.3
Wiring
A special cable for the Model DM8C Density Converter is connected to the Model VD6SM sampling
unit for the vibration type density meter.
2.3.1
Special Cable
The length of the special cable is as specified, and the cable end finishing (termination) is provided as
shown in Figure 2.7. When wiring, be careful not to stain or wet the cable end parts. If the cable is too
long, adjust the cable length by coiling the cable at the converter end, without cutting the cable.
Density converter
Density detector
B4
WHITE
B4
WHITE
GRAY B5
B5
GRAY
A2
RED
A2
RED
A3
BLUE
A3
BLUE
A4
YELLOW
A4
YELLOW
A5
GREEN
A5
GREEN
A6
BROWN
Figure 2.7
2.3.2
A6
BROWN
Special Cable
Wiiring Instructions
The wiring between the density detector and the density converter using the special cable is shown in
Figure 2.8.
Usually, the cable connected to the density detector is protected with a flexible fitting at the detector
cable inlet and from there to the cable duct. The cable is protected with a wire-reinforced-tube.
Detector
Special cable
Converter
A2
RED
A2
N.C
A3
BLUE
A3
C
A4
YELLOW
A4
A6
BROWN
A6
A5
GREEN
A5
B4
B5
WHITE
GRAY
SG
RD
TD
B5
L1 L2
+
+
-
Power supply
IM 12T3S1-01E
Wiring Diagram
RS-232-C signal
B4
G
Figure 2.8
Alarm
Closed when a
failure is detected or
the power fails.
Density signal after being
converted to the reference
temperature state
Density signal before being
converted to the reference
temperature state
7HPSHUDWXUHVLJQDO
0 to 1 V DC
4 to 20 mA DC
Density signal after being
converted to the reference
temperature state
< 2. INSTALLATION, PIPING AND WIRING >
2-7
Wiring for Non-explosionproof Type Density Detector
The cable inlet of the density detector is located at the bottom of the terminal box.
Through this cable inlet, the special cable is fed into the terminal box (remove the cover) and each
cable core wire should be connected to the correct terminals as shown in Figure 2.9. The terminal box
cover can be removed using the wrench supplied with the density detector.
Figure 2.9
Density Detector Terminal Connections
Figure 2.10
Wrench to Open/Close Terminal Box Cover
The conduit connecting screw at the cable inlet of the device is G3/4 Female. Use a proper flexible
fitting with this screw to protect the cable.
Cable
Terminal box case
Lock nut
Nut
G3/4 screw
Flaxible fitting
Figure 2.11
Cable Conduit Connections
IM 12T3S1-01E
2-8
< 2. INSTALLATION, PIPING AND WIRING >
Wiring for Explosionproof Density Detector
The explosionproof density detector conforms to either TIIS flameproof or FM explosionproof
specifications. The TIIS type is certified to be explosionproof by the Labor Safety and Sanitation Law
(Japan) which regulates domestic explosion-protected electrical equipment. The FM type is certified
to have an explosionproof construction meeting the requirements of NEC (National Electrical Code)
by FM (Factory Mutual Research Corporation, U.S.A.).
Use flameproof, metal conduit wiring for explosionproof density detectors. The conduit connecting
screw thread at the cable inlet of each device is either G3/4 female (TIIS explosionproof) or 3/4NPSM
(FM explosioprotection).
Use a proper explosionproof flexible fitting to protect the cable and to maintain the explosioproof
capability.
Cable
Terminal box case
Be sure to screw in
at least 12 mm (5 threads)
Lock nut
G3/4 screw (TIIS explosionproof)
3/4NPSM (FM explosionproof)
Explosionproof flaxible fitting
Figure 2.12
IM 12T3S1-01E
Explosionproof Flexible Fitting Connection
3-1
< 3. OPERATION >
3.
OPERATION
3.1
Names and Functions of Components
An external view of the sampling unit for the vibration type density meter with steam tracer is shown in
Figure 3.1. However, in the figure, the tracer pipe including heat insulating tape is omitted.
The valves “V1”, “V2” and the steam trap “ST1” are not provided with the general type sampling unit
for vibration type density meter.
Thermometer (T)
Pressure gauge (P)
Needle valve (NV2)
Ball valve (BV1)
Drain
Needle valve (NV3)
Density detector
Needle valve (NV4)
Drain
Pressure gauge
pressure element
Flowmeter (FM)
Valve (V1)
For density detector steam heating
Valve (V2)
For steam tracer in sampling unit piping
Needle valve (NV1)
Flow rate adjustment for measuring
liquid flowing through the density detector
Thermometer sensor
Steam trap (ST1)
Stariner (F)
Figure 3.1
Names and Functions of the Sampling Unit for Vibration Type Density Meter
(with Steam Tracer)
IM 12T3S1-01E
3-2
< 3. OPERATION >
3.2
3.2.1
Preparations for Operation
Removing Shipping Parts from Density Detector
To prevent possible damage during transportation, the vibrator assembly is separated from the
oscillation amplifier case by a rubber sheet. Before the unit is installed this rubber sheet should be
removed. The rubber sheet is inserted between the case and the vibrator assembly coupler. The
procedure is as follows:
(1)
Removing detector mainframe cover. First, loosen the lock screw with an Allen wrench (supplied
with the density detector) and remove the clamp. Turn the cover counterclockwise, then pull the
cover straight down, taking care that it does not bit the internal assembly.
(2)
Remove the vibrator assembly cover (contains desiccant).
NOTE
The vibrator is made of a pipe with very thin wall. Be careful not to dent or scratch it, otherwise it will
not function properly.
(3)
Loosen the vibrator lock screw. Use a flat blade screwdriver to turn the screw fully
counterclockwise; If the screw contacts the vibrator assembly during operation, invalid
measurements will result.
The vibrator is a pipe with thin wall. Hence, take care not to dent or damage the assembly.
(4)
Remove the rubber sheet. Pull it out, taking care not to apply excessive force to the vibrator
assembly.
Lock-screw
Use flat blade screw-driver
to turn this screw fully counterclockwise.
Pull rubber sheet
in this direction
Rubber sheet
Pull out this rubber sheet
after loosening the set screw.
Figure 3.2
(5)
Shipping Rubber Sheet Removal
Reassemble the cover removed in step (2), above. Then securely screw the cover removed in
step (1) above, onto the case assembly and lock it.
The removal of the rubber sheet is now completed. Save the removed rubber sheet: it is required
when shipping/transporting the density detector.
3.2.2
Inspecting Piping and Wiring Connections
Inspect the piping such as sample piping and all the wiring including the density converter or recorder
to confirm that they have been completed and they are in good condition.
If the density detector terminal box cover is removed for inspection, replace the cover after the
inspection and securely tighten it.
IM 12T3S1-01E
3-3
< 3. OPERATION >
3.3
Start-Up
In some cases, the density detector and the density converter – components of the vibration type
density meter – are installed at a distance from each other. Also, recorders and the like which receive
the density converter output signal may be installed at a distance from the density converter. This
section explains how to start up the unit.
However, in some descriptions, these components are assumed to be located close to each other, so
where necessary rewrite the procedure so that startup can be performed efficiently.
3.3.1
Connecting Power
Connect a power supply whose voltage and frequency are as specified for the density detector and
recorder for the output signal of the density detector. For recorders and the like which output control
signals, be careful that they do not cause abnormal operation.
3.3.2
Sampling Measurement Liquid
Pass the liquid to be measured through the sampling unit as follows:
(1)
Set the sampling unit values as follows:
Ball valve “BV1”:
Fully closed, Needle valve “NV1”:
Fully closed
Needle valve “NV2”: Fully closed, Needle valve “NV3”:
Fully closed
Needle valve “NV4”: Fully closed
*Stop valve “V1”:
Fully closed, *Stop valve “V2”: Fully closed
(Note) The valves V1 and V2 with * mark are only provided in the sampling unit with steam tracer.
T
b
P
b'
NV4
BV1
Sample
Inlet
FM
F
a
NV2
NV3
DD
NV1
Sample a'
Outlet
b
V1
a
V2
Steam
Inlet
b'
Condensate
Outlet
Figure 3.3
(2)
a'
ST1
T: Thermometer
BV1: Ball Valve
P: Pressure Gauge
NV1 to NV4: Needle Valve
F: Strainer
FM: Flow Meter
DD: Detector
V1, V2: StopValve
ST1: Steam Trap
Flow Diagram of the Sampling Unit for Vibration Type Density Meter (with
Steam Tracer)
When sampling lines are provided with steam tracer pipes, open the steam main valve to heat
the sample conduits. Fully open the stop valve “V2” to heat the piping in the sampling unit.
Leave the stop valve “V1” for heating the density detector fully closed unless the density detector
is very cold.
IM 12T3S1-01E
3-4
< 3. OPERATION >
(3)
Set a vessel to receive the liquid to be measured under the drain port provided for the sample
inlet pipe. Fully open the drain-valve, than tap valve. When the liquid to be measured flows out of
the drain, close the tap valve.
(4)
Carry out the same operation as described in (3) above for the sample outlet pipe.
(5)
Fully open the ball valve “BV1” and the needle valve “NV4”, half open the needle valve “NV1”,
and close the drain valve which was opened in (3) above. Gradually open the tap valve of the
sample inlet pipe. The liquid to be measured flows out of the drain port provided for the sample
outlet pipe. So that air is not mixed in the liquid to be measured, fully close the stop valve of this
piping and fully close the drain valve simultaneously.
(6)
To remove the remaining air in the piping, open the needle valve “NV2” and let the liquid to be
measured flow out of the drain port at the end of this valve. Then, close the valve.
(7)
Adjust the flow rate of the liquid to be measured flowing through the density detector, by
controlling the opening of the needle valve “NV1”. Normally, the flow rate is adjusted to
approximately 5 l/mm. The indication valve of the flowmeter “FM” is, as shown in Figure 3.4,
affected by the viscosity of the liquid to be measured, so for liquids of high viscosity the indicated
flow rate is higher than the actual flow rate.
$FWXDOIORZUDWHPHDVXUHPHQWOLTXLGGHQVLW\ȡ
F3
F3
F3
F3
F3
F3
F3
F3
F3
F3
F3
,QGLFDWLRQRIIORZPHWHUOPLQ
Figure 3.4
3.3.3
$FWXDOIORZUDWHOPLQ
Flow Rate Characteristics vs. Measurement Liquid Viscosity
(8)
Check to confirm that neither the thermometer “T1” nor the pressure gauge “P1” in the sampling
unit displays an abnormal valve (e.g. a valve exceeding the operating conditions).
(9)
Carry out visual inspection to confirm that there is no leakage of the liquid to be measured from
any pipe connection parts.
Calibration
To calibrate the vibration type density meter, see a separate User’s Manual IM 12T03A01-02E,
Section 4.4 “Calibration”.
IM 12T3S1-01E
3-5
< 3. OPERATION >
3.3.4
Measuring the Pressure Compensation Valves
When the calibration is carried out with standard solution, the pressure compensation valves should
be carried out as follows:
(1)
Fully open the needle valve “NV4” and read the pressure gauge “P1” indication valve.
(2)
Next, fully close the ball valve “BV1”, then open the needle valve “NV3” to set the pressure of the
liquid to be measured in the density detector to atmospheric pressure (0 MPa G).
Read the measured density valves under the two pressures described above (P1 and
atmospheric pressure) and obtain the pressure compensation valves as described in the other
user’s manual “Model DM8 Vibration type Density Meter ─ Pressure Compensation”.
3.3.5
Operation Check for Complete Density Measurement System
Operate the complete density measurement system under normal operating conditions.
Observe the system operation for a while to check for abnormal conditions.
The following items for must be checked:
(1) No measurement liquid leakage from piping connection.
(2) No air mixed in the measurement liquid and no disturbance in the measured valve.
(3) The temperature of the liquid to be measured is within the specified temperature range e.g. near
the reference temperature (within ±10 ºC).
(4) No large pressure variation of the liquid to be measured.
(5) The flow rate of the liquid to be measured is satisfactory and the necessary response time
required for the unit to operate well is obtained.
IM 12T3S1-01E
3-6
< 3. OPERATION >
3.4
Normal Operation
Normally, the sampling unit requires no adjustment during operation, however, to keep the unit in good
operating condition, daily check to confirm that the flow rate and pressure of the liquid to be measured
are normal.
The fo11owing instructions explain the inspection and maintenance to be carried out in normal
operation of the unit.
3.4.1
Instructions for Operation
No special care is required when the general use sampling unit for the vibration type density meter is
used. However, when the sampling unit for vibration type density meter with steam tracer is used, be
carefu1 not to heat the density detector. Always leave the stop valve “V1” fully closed.
The stop valve for the steam tracer piping in the sampling unit should be fully opened.
3.4.2
Cleaning Strainer Element
If the strainer element in the strainer assembly becomes clogged with dust, the flow rate of the liquid
to be measured flowing into the density detector decreases and the indication value of the flowmeter
“FM” decreases even though there has been no change in viscosity of the liquid to be measured. The
decrease in the flow rate can be somewhat compensated by opening the needle valve “NV1” a little,
however, ideally, the decreased flow rate due to clogging of the strainer element should be corrected
by cleaning the strainer element.
Clean the strainer element as follows:
(1)
Stop the flow of liquid to be measured.
First, fully close the tap valve provided for the sample inlet pipe. Next, fully close the stop valve
provided for the sample outlet pipe. If steam tracing is being carried out, stop the steam flow.
(2)
Drain the liquid to be measured from the sampling unit.
Set a vessel under the drain port provided for the sample inlet pipe, and open the drain valve of
this pipe and the needle valve “NV2” in the sampling unit.
(3)
Remove the strainer element.
First, fix the strainer body using a spanner and at the same time loosen the lower side nut of the
strainer. Take the gasket and strainer element out of the strainer assembly.
IM 12T3S1-01E
3-7
< 3. OPERATION >
Strainer
Element
Gasket
Nut of lower side
Strainer
to Flowmeter
Connector
Thermometer sensor
Figure 3.5
(4)
Strainer Assembly Construction
Clean the strainer element.
Using an appropriate solvent and taking care not to damage the mesh, remove the dust attached
to the strainer element.
3.4.3
(5)
Set the strainer element and gasket back in the strainer assembly, tighten the lower side nut of
the strainer.
(6)
Start the liquid to be measured flowing through the sampling unit again according to section
3.3.2.
Checking Measured Density Value
To get an idea of how often the unit will need to be calibrated or cleaned, it is recommended that you
check the measured density value as follows every few weeks.
(1)
Collect as sample of the liquid to be measured and measure its density with a standard
hydrometer.
When collecting the sample, confirm that the sample pressure and temperature are similar to
those during normal operation. Record the density reading of the vibration type density meter on
a recorder.
(2)
Accurately measure the density of the sample collected in (1) above.
(3)
Compare the density value measured in (2) above with the density value which was displayed on
the vibration type density meter and recorded in (1) above.
IM 12T3S1-01E
3-8
< 3. OPERATION >
If the difference between these values exceeds the maximum allowable error, clean the vibrator
in the density detector.
3.4.4
Cleaning Vibrator in the Density Detector
Clean the vibrator as fo11ows:
(1)
Close the ball valve “BV1” and the needle valve “NV4” in the sampling unit, and open the needle
valves “NV2” and “NV3”.
(2)
Clean the vibrator referring to the other instruction manual “Model DM8 vibration type liquid
density measuring system” section 5.1.1 “Cleaning Vibrator”.
(3)
Close the needle valves “NV2” and “NV3” and fully open the ball valve “BV1” and the needle
valve “NV4” to start the liquid to be measured flowing through the density detector.
Purge the remaining air in the piping from the drain on the needle valve “NV2” side.
(4)
3.4.5
Check the measured density value of the vibration type density meter according to section
3.4.3. If the difference between the measured values of the vibration type density meter and the
measured value of the standard hydrometer exceeds the allowable range, recalibrate the unit.
Cautions When Stopping the Operation
When stopping flow of the liquid to be measured and when the sampling unit is not in operation, as a
rule, drain the measurement liquid as follows.
(1)
Close the tap valve provided for the sample inlet pipe and the stop valve provided for the
sample outlet pipe. Next close the drain valves provided for these pipes. A part of the liquid to be
measured in the sampling unit will flow out of each drain valve. If required, a vessel to receive the
liquid to be measured should be prepared under each drain port.
(2)
Drain the liquid to be measured in the sampling unit as far as possible. First, open the needle
valve “NV2”. Next, fully close the ball valve “BV1”, connect the pneumatic piping to the drain
port at the needle valve “NV2” side, and apply a pressure of 0.1 to 0.2 MPa of air to the unit for
several minutes.
(3)
If the liquid to be measured may stick to the inside of the vibrator, clean the vibrator. If required,
inject an appropriate solvent into the vibrator.
(4)
Fully close the drain valves of the sample conduits. The needle valve “NV2” should be slightly
open.
When the sampling unit with steam tracer is used, always full close the stop valves “V1” and
“V2”’.
3.4.6
Consumable Part Replacement
The Model VD6SM Sampling unit for vibration type density meter contains no consumable parts
which must be replaced periodically. For the strainer assembly, which may have to be comparatively
frequently replaced, a strainer element and on gasket are provided as spare parts with the unit, so
when cleaning the unit carry out visual inspection and if they are damaged, replace them.
Should any failure occur during long term usage, replace the failed part as soon as possible.
IM 12T3S1-01E
Customer
Maintenance
Parts List
Model VD6SM*B
Sampling Unit
for Vibration Type Density Meter
1
2
3
4
10
5
6
5
7
5
8
9
9
8
11
12
©Copyright Aug. 2005, 1st Edition: Aug, 2005 (YK)
Subject to change without notice.
CMPL 12T03S01-02E
2
2
3
4
5
Part No.
L9868AP
L9868AQ
Y9512JU
Y9500WU
Y9501BU
L9852CB
6
7
8
9
L9854BX
–
L9866BV
L9852CA
10
11
12
1
1
6
12
6
4
1
1
6
12
6
4
Description
Thermometer (0 to 1008C)
Thermometer (0 to 1508C)
Screw
Washer
Nut
Needle Valve
1
1
1
1
1
1
2
Ball Valve
Density Detector (see CMPL 12T3E1-01E)
Flowmeter
Stop Valve
1
1
Pressure Gauge
0 to 1 MPa
0 to 2 MPa
0 to 1 MPa
0 to 2 MPa
–
1
L9867BV 1
L9867BW 1
L9867BX
L9867BY
L9865AE
–
K9727KK
K9727LG
Note *1
Note *2
Note *2
Item
1
Model
Note *1
Qty
1
1
1
1
1
Steam Trap
Strainer Assembly (for details see below)
with SUS316 element
with Ni element (option code: /FN)
Model VD6SM - uuu - Puu0 - T1u0 *B
Model VD6SM - uuu - Puu0 - T1u0 *B /ST
Item 12 Strainer Assembly
1
2
Qty
*1 *2
K9727KX 1
K9727LC
1
K9727KW 1 1
Item Part Name Part No.
1
Element
2
Gasket
Description
SUS316
Ni
Teflon TM
(Note) *1 Model VD6SM - uuu -Puu0- T1u0 *B
*2 Model VD6SM - uuu -Puu0- T1u0 *B /FN
1st Edition: Aug, 2005 (YK)
CMPL 12T03S01-02E
i
Revision Information
Title
Manual No.
: Model VD6SM Sampling Unit For Vibration Type Density Meter
: IM 12T3S1-01E
Jul. 2012/3rd Edition Page layout changed by InDesign
P. 1-1 Standard specifications equalized to GS; P. 2-3 Pressure value of Figure 2.2 modified, etc.
Oct. 2006/2nd Edition
Style change from *A to *B.
Mar. 1988/1st Edition
Newly published
If you want to have more information about Yokogawa products, you can visit
Yokogawa’s home page at the following web site.
Home page: http://www.yokogawa.com/an
IM 12T3S1-01E
Blank Page