Emerson 1056 DUAL-INPUT INTELLIGENT ANALYZER Instruction Manual
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Instruction Manual PN 51­1056/rev.J May 2014 DUAL-INPUT INTELLIGENT ANALYZER 1056 ESSENTIAL INSTRUCTIONS WARNING READ THIS PAGE BEFORE PROCEEDING! Your instrument purchase from Rosemount Analytical, Inc. is one of the finest available for your particular application. These instruments have been designed, and tested to meet many national and international standards. Experience indicates that its performance is directly related to the quality of the installation and knowledge of the user in operating and maintaining the instrument. To ensure their con­ tinued operation to the design specifications, per­ sonnel should read this manual thoroughly before proceeding with installation, commissioning, opera­ tion, and maintenance of this instrument. If this equipment is used in a manner not specified by the manufacturer, the protection provided by it against hazards may be impaired. • Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property dam­ age; damage to this instrument; and warranty invalidation. • Ensure that you have received the correct model and options from your purchase order. Verify that this manual covers your model and options. If not, call 1­800­854­8257 or 949­757­8500 to request correct manual. • For clarification of instructions, contact your Rosemount representative. RISK OF ELECTRICAL SHOCK Equipment protected throughout by double insulation. • Installation and servicing of this product may expose personel to dangerous voltages. • Main power wired to separate power source must be disconnected before servicing. • Do not operate or energize instrument with case open! • Signal wiring connected in this box must be rated at least 240 V. • Non­metallic cable strain reliefs do not provide grounding between conduit connections! Use grounding type bushings and jumper wires. • Unused cable conduit entries must be securely sealed by non­flammable closures to provide enclosure integrity in compliance with personal safety and environmental protection requirements. Unused conduit openings must be sealed with NEMA 4X or IP65 conduit plugs to maintain the ingress protection rating (NEMA 4X). • Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA­70) and/or any other applicable national or local codes. • Operate only with front panel fastened and in place. • Safety and performance require that this instrument be connected and properly grounded through a three­wire power source. • Proper use and configuration is the responsibility of the user. • Follow all warnings, cautions, and instructions marked on and supplied with the product. • Use only qualified personnel to install, operate, update, program and maintain the product. • Educate your personnel in the proper installation, operation, and maintenance of the product. • Install equipment as specified in the Installation section of this manual. Follow appropriate local and national codes. Only connect the product to electrical and pressure sources specified in this manual. • Use only factory documented components for repair. Tampering or unauthorized substitution of parts and procedures can affect the performance and cause unsafe operation of your process. • All equipment doors must be closed and protec­ tive covers must be in place unless qualified per­ sonnel are performing maintenance. Emerson Process Management 2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757­8500 Fax: (949) 474­7250 http://www.rosemountanalytical.com © Rosemount Analytical Inc. 2012 CAUTION This product generates, uses, and can radiate radio frequency energy and thus can cause radio communication interference. Improper installation, or operation, may increase such interfer­ ence. As temporarily permitted by regulation, this unit has not been tested for compliance within the limits of Class A comput­ ing devices, pursuant to Subpart J of Part 15, of FCC Rules, which are designed to provide reasonable protection against such interference. Operation of this equipment in a residential area may cause interference, in which case the user at his own expense, will be required to take whatever measures may be required to correct the interference. CAUTION This product is not intended for use in the light industrial, residential or commercial environments per the instru­ ment’s certification to EN50081­2. QUICK START GUIDE 1056 Dual Input Analyzer 1. Refer to Section 2.0 for mechanical installation instructions. 2. Wire sensor(s) to the signal boards. See Section 3.0 for wiring instructions. Refer to the sensor instruction sheet for additional details. Make current output, alarm relay and power connections. 3. Once connections are secured and verified, apply power to the analyzer. WARNING RISK OF ELECTRICAL SHOCK Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA­70) and/or any other applicable national or local codes. 4. When the analyzer is powered up for the first time, Quick Start screens appear. Quick Start operating tips are as follows: a. A backlit field shows the position of the cursor. b. To move the cursor left or right, use the keys to the left or right of the ENTER key. To scroll up or down or to increase or decrease the value of a digit use the keys above and below the ENTER key . Use the left or right keys to move the decimal point. c. Press ENTER to store a setting. Press EXIT to leave without storing changes. Pressing EXIT during Quick Start returns the display to the initial start­up screen (select language). 5. Complete the steps as shown in the Quick Start Guide flow diagram, Fig. A on the following page. 6. After the last step, the main display appears. The outputs are assigned to default values. 7. To change output, and temperature­related settings, go to the main menu and choose Program. Follow the prompts. For a general guide to the Program menu, see the Quick Reference Guide, Fig.B. 8. To return the analyzer to the default settings, choose Reset Analyzer under the Program menu. Figure A. QUICK START GUIDE QUICK START GUIDE Figure B. MODEL 1056 MENU TREE QUICK REFERENCE GUIDE About This Document This manual contains instructions for installation and operation of the Model 1056 Dual­Input Intelligent Analyzer. The following list provides notes concerning all revisions of this document. Rev. Level Date Notes A 01/07 B C 2/07 9/07 D 11/07 E F G H I J 05/08 08/08 09/08 04/10 03/12 05/14 This is the initial release of the product manual. The manual has been reformatted to reflect the Emerson documentation style and updated to reflect any changes in the product offering. Added CE mark to p.2. Replaced Quick Start Fig A. Revised Sections 1,3,5,6, and 7. Added new measurements and features ­ Turbidity, Flow, Current Input, Alarm relays and 4­electrode conductivity. Added 24VDC power supply to Sec. 3.4. Added CSA and FM agency approvals for option codes ­01, 20, 21, 22, 24, 25, 26, 30, 31, 32, 34, 35, 36 and 38. Add HART and Profibus DP digital communication to Section 1 specifications. Updates FM and CSA agency approval, Class 1, Div 2. for 24 VDC and AC switching power supplies. Update DNV logo and company name Update addresses ­ mail and web Update enclosure information MODEL 1056 TABLE OF CONTENTS MODEL 1056 DUAL INPUT INTELLIGENT ANALYZER TABLE OF CONTENTS QUICK START GUIDE QUICK REFERENCE GUIDE TABLE OF CONTENTS Section Title 1.0 DESCRIPTION AND SPECIFICATIONS ................................................................ 2.0 INSTALLATION ....................................................................................................... 2.1 Unpacking and Inspection........................................................................................ 2.2 Installation................................................................................................................ Page 1 11 11 11 3.0 3.1 3.2 3.3 3.4 WIRING.................................................................................................................... General .................................................................................................................... Preparing Conduit Openings.................................................................................... Preparing Sensor Cable .......................................................................................... Power, Output, Alarms and Sensor Connections..................................................... 19 19 19 20 20 4.0 4.1 4.2 4.3 4.4 5.0 DISPLAY AND OPERATION ................................................................................... User Interface .......................................................................................................... Instrument Keypad................................................................................................... Main Display ............................................................................................................ Menu System ........................................................................................................... PROGRAMMING – BASICS ................................................................................... 27 27 27 28 29 31 5.1 5.2 5.3 5.4 5.5 5.6 5.7 5.8 5.9 6.0 General .................................................................................................................... Changing the StartUp Settings ................................................................................ Choosing Temperature units and Automatic/Manual Temperature Compensation .. Configuring and Ranging the Current Outputs......................................................... Setting a Security Code ........................................................................................... Security Access........................................................................................................ Using Hold ............................................................................................................... Resetting Factory Defaults – Reset Analyzer .......................................................... Alarm Relays............................................................................................................ PROGRAMMING ­ MEASUREMENTS ................................................................... 31 31 32 32 34 35 35 36 37 41 6.1 6.2 6.3 6.4 6.5 6.6 Programming Measurements – Introduction ........................................................... pH ............................................................................................................................ ORP ......................................................................................................................... Contacting Conductivity .......................................................................................... Toroidal Conductivity................................................................................................ Chlorine.................................................................................................................... 41 42 43 45 48 51 6.6.1 Free Chlorine .................................................................................................. 51 6.6.2 Total Chlorine ................................................................................................. 53 6.6.3 Monochloramine ............................................................................................ 54 6.6.4 pH­independent Free Chlorine ....................................................................... 55 Oxygen..................................................................................................................... Ozone ...................................................................................................................... 57 59 6.7 6.8 i MODEL 1056 TABLE OF CONTENTS TABLE OF CONTENTS CONT’D 6.9 Turbidity ................................................................................................................... 60 6.10 Flow ......................................................................................................................... 63 6.11 Current Input ............................................................................................................ 64 7.0 7.1 CALIBRATION ...................................................................................................... Calibration – Introduction ......................................................................................... 75 75 7.2 pH Calibration .......................................................................................................... 76 7.3 ORP Calibration ....................................................................................................... 78 7.4 Contacting Conductivity Calibration ......................................................................... 79 7.5 Toroidal Conductivity Calibration ............................................................................. 82 7.6 Chlorine Calibration ................................................................................................. 84 7.6.1 Free Chlorine .................................................................................................. 84 7.6.2 Total Chlorine .................................................................................................. 86 7.6.3 Monochloramine ............................................................................................. 88 7.6.4 pH­Independent Free Chlorine ....................................................................... 90 7.7 Oxygen Calibration .................................................................................................. 92 7.8 Ozone Calibration .................................................................................................... 95 7.9 Temperature Calibration........................................................................................... 97 7.10 Turbidity ................................................................................................................... 98 7.11 Pulse Flow ............................................................................................................... 100 8.0 RETURN OF MATERIAL ........................................................................................ Warranty................................................................................................................... 112 112 ii MODEL 1056 TABLE OF CONTENTS TABLE OF CONTENTS CONT’D LIST OF FIGURES Fig# A B 2­1 2­2 2­3 2­4 2­5 2­6 3­1 3­2 3­3 3­4 3­5 3­6 3­7 3­8 3­9 3­10 3­11 3­12 3­13 3­14 3­15 4­1 5­1 5­2 5­3 5­4 5­5 6­1 6­2 6­3 6­4 6­5 6­6 6­7 6­8 6­9 7­1 7­2 7­3 7­4 7­5 7­6 7­7 7­8 7­9 Section PREFACE PREFACE SEC 2.0 SEC 2.0 SEC 2.0 SEC 2.0 SEC 2.0 SEC 2.0 SEC 3.4 SEC 3.4 SEC 3.4 SEC 3.4 SEC 3.4 SEC 3.4 SEC 3.4 SEC 3.4 SEC 3.4 SEC 3.4 SEC 3.4 SEC 3.4 SEC 3.4 SEC 3.4 SEC 3.4 SEC 4.3 SEC 5.3.2 SEC 5.4.5 SEC 5.5.2 SEC 5.7.2 SEC 5.8.2 SEC 6.2 SEC 6.4 SEC 6.5 SEC 6.6 SEC 6.7 SEC 6.8 SEC 6.9 SEC 6.10 SEC 6.11 SEC 7.2 SEC 7.3 SEC 7.4 SEC 7.6 SEC 7.7 SEC 7.8 SEC 7.9 SEC 7.10 SEC 7.11 Figure Title Quick Start Guide Quick Reference Guide Panel Mounting Dimensions .............................................................. Pipe and Wall Mounting Dimensions ................................................. CSA Certification drawing part 1 ........................................................ CSA Certification drawing part 2 ........................................................ FM Non­Incendive drawing part 1...................................................... FM Non­Incendive drawing part 2...................................................... 115/230 VAC Power Supply ............................................................... 24VDC Power Supply ........................................................................ Switching AC Power Supply............................................................... Current Output Wiring ........................................................................ Alarm Relay Wiring for Model 1056 Switching Power Supply............ Contacting Conductivity board and sensor cable leads ..................... Toroidal Conductivity signal board and sensor cable leads ............... pH/ORP/ISE signal board and sensor cable leads ............................ Amperometric board (Cl, O2, Ozone) and sensor cable leads .......... Turbidity signal board wiht plug­in Sensor connection....................... Flow/Current Input signal board and Sensor cable leads .................. Power Wiring for Model 1056 115/230 VAC....................................... Power Wiring for Model 1056 85­265 VAC ........................................ Output Wiring for Model 1056 Main PCB........................................... Power Wiring for Model 1056 24VDC ................................................ Formatting the Main Display ............................................................. Choosing Temp Units and Manual Auto Temp Compensation ........... Configuring and Ranging the Current Outputs................................... Setting A Security Code .................................................................... Using Hold ......................................................................................... Resetting Factory Default Settings .................................................... Configuring pH/ORP Measurements ................................................. Configure Contacting Measurements ............................................... Configure Toroidal Measurements .................................................... Configure Chlorine Measurements .................................................... Configure Oxygen Measurements ..................................................... Configure Ozone Measurements ....................................................... Configure Turbidity Measurement...................................................... Configure Flow Measurement............................................................ Configure mA Current Input Measurement ........................................ Calibrate pH ....................................................................................... Calibrate ORP.................................................................................... Calibrate Contacting and Toroidal Conductivity ................................. Calibrate Chlorine .............................................................................. Calibrate Oxygen ............................................................................... Calibrate Ozone ................................................................................. Calibrate Temperature ....................................................................... Calibrate Turbidity .............................................................................. Calibrate Flow .................................................................................... Page 12 13 14 15 16 17 20 20 20 21 21 22 22 23 23 24 24 25 25 26 26 30 32 33 34 35 36 67 68 69 71 70 71 72 73 73 103 104 105 106 107 108 109 110 111 iii MODEL 1056 TABLE OF CONTENTS TABLE OF CONTENTS CONT’D LIST OF TABLES Number Section iv Table Title Page 5­1 SEC 5.2.1 Measurements and Measurement Units ............................................. 31 6­1 SEC 6.2.1 pH Measurement Programming ........................................................... 42 6­2 SEC 6.3.1 ORP Measurement Programming ........................................................ 43 6­3 SEC 6.4.1 Contacting Conductivity Measurement Programming.......................... 45 6­4 SEC 6.5.1 Toroidal Conductivity Measurement Programming............................... 48 6­5 SEC 6.6.1.1 Free Chlorine Measurement Programming.......................................... 51 6­6 SEC 6.6.2.1 Total Chlorine Measurement Programming ......................................... 53 6­7 6­8 6­9 6­10 6­11 6­12 6­13 7­1 7­2 7­3 7­4 7­5 7­6 7­7 7­8 7­9 7­10 7­11 7­12 7­13 SEC 6.6.3.1Monochloramine Measurement Programming..................................... SEC 6.6.4 pH­Independent Free Chlorine Measurement Programming ............ SEC. 6.7.1 Oxygen Measurement Programming.................................................... SEC 6.8.1 Ozone Measurement Programming...................................................... SEC 6.9.1 Turbidity Measurement Programming.................................................. SEC 6.10.1 Flow Measurement Programming......................................................... SEC 6.11.1 Curent Input Programming................................................................... SEC 7.2 pH Calibration Routines ....................................................................... SEC 7.3 ORP Calibration Routine...................................................................... SEC 7.4 Contacting Conductivity Calibration Routines...................................... SEC 7.5 Toroidal Conductivity Calibration ........................................... ............ SEC 7.6.1 Free Chlorine Calibration Routines....................................................... SEC 7.6.2 Total Chlorine Calibration Routines...................................................... SEC 7.6.3 Monochloramine Calibration Routines.................................................. SEC 7.6.4 pH­independent Free Chlorine Calibration Routines............................ SEC 7.7 Oxygen Calibration Routines................................................................ SEC 7.8 Ozone Calibration Routines.................................................................. SEC 7.9 Temperature Calibration Routines........................................................ SEC 7.10 Turbidity Calibration Routines............................................................... SEC 7.11 Flow Calibration Routines................................................................... 54 55 57 59 60 63 64 76 78 79 82 85 86 88 90 93 95 97 98 100 MODEL 1056 SECTION 1.0 DESCRIPTION AND SPECIFICATIONS SECTION 1.0. DESCRIPTION AND SPECIFICATIONS • MULTI­PARAMETER INSTRUMENT – single or dual input. Choose from pH/ORP/ISE, Resistivity/Conductivity, % Concentration, Chlorine, Oxygen, Ozone, Temperature, Turbidity, Flow, and 4­20mA Current Input. • LARGE DISPLAY – large easy­to­read process measurements. • EASY TO INSTALL – modular boards, removable connectors, easy to wire power, sensors, and outputs. • INTUITIVE MENU SCREENS with advanced diagnostics and help screens. • SEVEN LANGUAGES included: English, French, German, Italian, Spanish, Portuguese, and Chinese. • HART® AND PROFIBUS® DP Digital Communications options FEATURES AND APPLICATIONS The 1056 dual­input analyzer offers single or dual sen­ sor input with an unrestricted choice of dual measure­ ments. This multi­parameter instrument offers a wide range of measurement choices supporting most indus­ trial, commercial, and municipal applications. The modular design allows signal input boards to be field replaced making configuration changes easy. Conveniently, live process values are always displayed during programming and calibration routines. QUICK START PROGRAMMING: Exclusive Quick Start screens appear the first time the 1056 is pow­ ered. The instrument auto­recognizes each measure­ ment board and prompts the user to configure each sensor loop in a few quick steps for immediate deploy­ ment. DIGITAL COMMUNICATIONS: HART and Profibus DP digital communications are available. The 1056 HART units communicate with the Model 375 HART® hand­held communicator and HART hosts, such as AMS Intelligent Device Manager. Model 1056 Profibus units are fully compatible with Profibus DP networks and Class 1 or Class 2 masters. HART and Profibus DP configured units will support any single or dual measurement configuration of Model 1056. MENUS: Menu screens for calibrating and programming are simple and intuitive. Plain language prompts and help screens guide the user through these procedures. DUAL SENSOR INPUT AND OUTPUT: The 1056 accepts single or dual sensor input. Standard 0/4­ 20 mA current outputs can be programmed to correspond to any measurement or temperature. ENCLOSURE: The instrument fits standard ½ DIN panel cutouts. The versatile enclosure design supports panel­mount, pipe­mount, and surface/wall­mount installations. ISOLATED INPUTS: Inputs are isolated from other signal sources and earth ground. This ensures clean signal inputs for single and dual input configurations. For dual input configurations, isolation allows any combination of measurements and signal inputs with­ out cross­talk or signal interference. TEMPERATURE: Most measurements require tem­ perature compensation. The 1056 will automatically recognize Pt100, Pt1000 or 22k NTC RTDs built into the sensor. SECURITY ACCESS CODES: Two levels of security access are available. Program one access code for routine calibration and hold of current outputs; program another access code for all menus and functions. 1 MODEL 1056 SECTION 1.0 DESCRIPTION AND SPECIFICATIONS DIAGNOSTICS: The analyzer continuously monitors itself and the sensor(s) for problematic conditions. The display flashes Fault and/or Warning when these conditions occur. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC Diagnostics Faults Warnings Sensor 1 Sensor 2 Out 1: 12.05 mA Out 2: 12.05 mA 1056­01­20­32­HT Instr SW VER: 2.12 AC Freq. Used: 60Hz Information about each condition is quickly accessible by pressing DIAG on the keypad. User help screens are displayed for most fault and warning conditions to assist in troubleshooting. DISPLAY: The high­contrast LCD provides live measurement readouts in large digits and shows up to four additional process variables or diagnostic parameters. The display is back­lit and the format can be customized to meet user requirements. LOCAL LANGUAGES : Rosemount Analytical extends its worldwide reach by offering seven local languages – English, French, German, Italian, Spanish, Portuguese, and Chinese. Every unit includes user programming menus; calibration routines; faults and warnings; and user help screens in all seven languages. The displayed language can be easily set and changed using the menus. CURRENT OUTPUTS: Two 4­20 mA or 0­20 mA current outputs are electrically isolated. Outputs are fully scalable and can be programmed to linear or logarithmic modes. Output dampening can be enabled with time constants from 0 to 999 seconds. Output 1 includes digital signal 4­20 mA superimposed HART (option ­HT only) 2 SPECIAL MEASUREMENTS: The Model 1056 offers measuring capabilities for many applications. l Single or Dual Turbidity: Ideal in municipal appli­ cations for measurement of low­NTU filtered drinking water. Must be used with Clarity II sensor, sensor cable and debubbler. Model T1056 Clarity® II Turbidimeter System l 4­Electrode Conductivity: The 1056 is compatible with Rosemount Analytical 4­electrode Model 410VP in the PUR­SENSE family of conductivity sensors. This sensor supports a wide array of applications and is capable of measuring a large range of conductivity with one geometric configuration. Wired to the 1056, this sensor can measure 2μS/cm to 300mS/cm with an accuracy of 4% of reading throughout the entire range. l 4­20mA Current Input: Accepts any analog current input from an external device for temperature compen­ sation of measurements and atmospheric pressure input for partial pressure correction of oxygen. l Selective Ions: The analyzer is able to measure ammonia and fluoride using commercially available ion­selective electrodes. All analyzers with installed pH boards can be programmed to measure selective ions. l pH Independent Free Chlorine: With Rosemount Analytical’s 498Cl­01 sensor, the analyzer is able to measure free chlorine with automatic correction for process pH without the need for a pH sensor. l Inferential pH: The analyzer is able to derive and display inferred pH (pHCalc) using two contacting con­ ductivity signal boards and the appropriate contacting conductivity sensors. This method will calculate the pH of condensate and boiler water from conductivity and cation conductivity measurements. l Differential Conductivity: Dual input conductivity configurations can measure differential conductivity. The analyzer can be programmed to display dual conductivity as ratio, % rejection, or % passage. MODEL 1056 SECTION 1.0 DESCRIPTION AND SPECIFICATIONS SPECIFICATIONS ­ General Enclosure: Polycarbonate. Type 4X, IP65. Note: To ensure a water­tight seal, tighten all four front panel screws to 6 in­lbs of torque Power: Code ­01: 115/230 VAC ±15%, 50/60 Hz. 10W. Code ­02: 20 to 30 VDC. 15 W. Code ­03: 85 to 265 VAC, 47.5 to 65.0 Hz, switching. 15 W. Note: Code ­02 and ­03 power supplies include 4 pro­ grammable relays Equipment protected by double insulation RFI/EMI: EN­61326 LVD: EN­61010­1 Alarms relays*: Four alarm relays for process meas­ urement(s) or temperature. Any relay can be config­ ured as a fault alarm instead of a process alarm. Each relay can be configured independently and each can be programmed with interval timer settings. Relays: Form C, SPDT, epoxy sealed Dimensions: Overall 155 x 155 x 131mm (6.10 x 6.10 x 5.15 in.). Cutout: 1/2 DIN 139mm x 139mm (5.45 x 5.45 in.) Conduit Openings: Accepts 1/2” or PG13.5 conduit fittings Display: Monochromatic graphic liquid crystal display. 128 x 96 pixel display resolution. Backlit. Active display area: 58 x 78mm (2.3 x 3.0 in.). Ambient Temperature and Humidity: 0 to 55°C (32 to 131°F). Turbidity only: 0 to 50°C (32 to 122°F), RH 5 to 95% (non­condensing) Hazardous Location Approvals ­ Options for CSA: ­01, 02, 03, 20, 21, 22, 24, 25, 26, 27, 30, 31, 32, 34, 35, 36, 37, 38, AN, and HT. Class I, Division 2, Groups A, B, C, & D Class Il, Division 2, Groups E, F, & G Class Ill T4A Tamb= 50° C Evaluated to the ANSI/UL Standards. The ‘C’ and ‘US’ indi­ cators adjacent to the CSA Mark signify that the product has been evaluated to the applicable CSA and ANSI/UL Standards, for use in Canada and the U.S. respectively Options for FM: ­01, 02, 03, 20, 21, 22, 24, 25, 26, 30, 31, 32, 34, 35, 36, 38, AN, and HT. Class I, Division 2, Groups A, B, C, & D Class Il & lll, Division 2, Groups E, F, & G T4A Tamb= 50° C Enclosure Type 4X Storage Temperature Effect: ­20 to 60ºC (­4 to 140°F) POLLUTION DEGREE 2: Normally only non­conductive pollution occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected. Altitude: for use up to 2000 meter (6562 ft.) Maximum Relay Current Resistive 28 VDC 115 VAC 230 VAC 5.0 A 5.0 A 5.0 A Inductive load: 1/8 HP motor (max.), 40 VAC CAUTION RISK OF ELECTRICAL SHOCK *Relays only available with ­02 power supply (20 ­ 30 VDC) or ­03 switching power supply (85 ­ 265 VAC) WARNING WARNING Exposure to some chemicals may degrade the sealing properties used in the following devices: Zettler Relays (K1­K4) PN AZ8­1CH­12DSEA Inputs: One or two isolated sensor inputs Outputs: Two 4­20 mA or 0­20 mA isolated current out­ puts. Fully scalable. Max Load: 550 Ohm. Output 1 has superimposed HART signal (configurations 1056­0X­2X­3X­HT only) Current Output Accuracy: ±0.05 mA @ 25 ºC Terminal Connections Rating: Power connector (3­leads): 24­12 AWG wire size. Signal board ter­ minal blocks: 26­16 AWG wire size. Current output connectors (2­leads): 24­16 AWG wire size. Alarm relay terminal blocks: 24­12 AWG wire size (­02 24 VDC power supply and ­03 85­265VAC power supply) Weight/Shipping Weight: (rounded up to nearest lb or nearest 0.5 kg): 3 lbs/4 lbs (1.5 kg/2.0 kg) 3 MODEL 1056 SECTION 1.0 DESCRIPTION AND SPECIFICATIONS CONTACTING CONDUCTIVITY (Codes ­20 and ­30) Measures conductivity in the range 0 to 600,000 μS/cm (600mS/cm). Measurement choices are conductivity, resistivity, total dissolved solids, salinity, and % concen­ tration. The % concentration selection includes the choice of five common solutions (0­12% NaOH, 0­15% HCl, 0­20% NaCl, and 0­25% or 96­99.7% H2SO4). The conductivity concentration algorithms for these solutions are fully temperature compensated. Three temperature compensation options are available: manual slope (X%/°C), high purity water (dilute sodium chloride), and cation conductivity (dilute hydrochloric acid). Temperature compensation can be disabled, allowing the analyzer to display raw conductivity. For more information concerning the use and operation of the contacting conductivity sensors, refer to the product data sheets. Temperature Specifications: Temperature range 0­150ºC Temperature Accuracy, Pt­1000, 0­50 ºC ± 0.1ºC Temperature Accuracy, Pt­1000, Temp. > 50 ºC ± 0.5ºC RECOMMENDED SENSORS FOR CONDUCTIVITY: All Rosemount Analytical ENDURANCE Model 400 series conductivity sensors (Pt 1000 RTD) and Model 410 sensor. Note: When two contacting conductivity sensors are used, Model 1056 can derive an inferred pH value called pHCalc. pHCalc is calculated pH, not directly measured pH. (Model 1056­0X­20­30­AN required) Note: Selected 4­electrode, high­range contacting conductivity sensors are compatible with Model 1056. Input filter: time constant 1 ­ 999 sec, default 2 sec. family 4­electrode sensors Response time: 3 seconds to 100% of final reading Salinity: uses Practical Salinity Scale Total Dissolved Solids: Calculated by multiplying conductivity at 25ºC by 0.65 ENDURANCE series of conductivity sensors TM PERFORMANCE SPECIFICATIONS Recommended Range – Contacting Conductivity Cell Constant 0.01 0.1 1.0 4­electrode 0.01μS/cm 0.1μS/cm 1.0μS/cm 10μS/cm 100μS/cm 0.01μS/cm to 200μS/cm 1000μS/cm 10mS/cm 100mS/cm 200μS/cm to 6000μS/cm 2000μS/cm to 60mS/cm 0.1μS/cm to 2000μS/cm 1 μS/cm to 20mS/cm 20mS/cm to 600mS/cm 2 μS/cm to 300mS/cm Cell Constant Linearity ±0.6% of reading in recommended range +2 to ­10% of reading outside high recommended range 4 1000mS/cm ±5% of reading outside low recommended range ±4% of reading in recommended range MODEL 1056 SECTION 1.0 DESCRIPTION AND SPECIFICATIONS TOROIDAL CONDUCTIVITY (Codes ­21 and ­31) Measures conductivity in the range of 1 (one) μS/cm to 2,000,000 μS/cm (2 S/cm), Measurement choices are conductivity, resistivity, total dissolved solids, salinity, and % concentration. The % concentration selection includes the choice of five common solutions (0­12% NaOH, 0­15% HCl, 0­20% NaCl, and 0­25% or 96­99.7% H2SO4). The conductivity concentration algorithms for these solutions are fully temperature compensated. For other solutions, a simple­to­use menu allows the customer to enter his own data. The analyzer accepts as many as five data points and fits either a linear (two points) or a quadratic function (three or more points) to the data. Two temperature compensation options are available: manual slope (X%/°C) and neutral salt (dilute sodium chloride). Temperature compensation can be disabled, allowing the analyzer to display raw conductivity. Reference temperature and linear temper­ ature slope may also be adjusted for optimum results. For more information concerning the use and operation of the toroidal conductivity sensors, refer to the product data sheets. Temperature Specifications: Temperature range ­25 to 210ºC (­13 to 410ºF) Temperature Accuracy, Pt­100, ­25 to 50 ºC ± 0.5ºC Temperature Accuracy, Pt­100,. 50 to 210ºC ± 1ºC RECOMMENDED SENSORS: All Rosemount Analytical submersion/immersion and flow­through toroidal sensors. Repeatability: ±0.25% ±5 μS/cm after zero cal Input filter: time constant 1 ­ 999 sec, default 2 sec. Response time: 3 seconds to 100% of final reading Salinity: uses Practical Salinity Scale Total Dissolved Solids: Calculated by multiplying conductivity at 25ºC by 0.65 High performance toroidal conductivity sensors Models 226 and 225 PERFORMANCE SPECIFICATIONS Recommended Range ­ Toroidal Conductivity Model 226 1μS/cm 10μS/cm 100μS/cm 1000μS/cm 10mS/cm 100mS/cm 5μS/cm to 500mS/cm 1000mS/cm 2000mS/cm 500mS/cm to 2000mS/cm 225 & 228 15μS/cm to 1500mS/cm 242 222 (1in & 2in) 1500mS/cm to 2000mS/cm 100μS/cm to 2000mS/cm 500μS/cm to 2000mS/cm LOOP PERFORMANCE (Following Calibration) Model 226: ±1% of reading ±5μS/cm in recommended range Models 225 & 228: ±1% of reading ±10μS/cm in recommended range Models 222,242: ±4% of reading in recommended range Model 225, 226 & 228: ±5% of reading outside high recommended range Model 226: ±5μS/cm outside low recommended range Models 225 & 228: ±15μS/cm outside low recommended range 5 MODEL 1056 SECTION 1.0 DESCRIPTION AND SPECIFICATIONS pH/ORP/ISE (Codes ­22 and ­32) For use with any standard pH or ORP sensor. Measurement choices are pH, ORP, Redox, ammonia, fluoride or custom ISE. The automatic buffer recognition feature uses stored buffer values and their temperature curves for the most common buffer standards available worldwide. The analyzer will recognize the value of the buffer being measured and perform a self stabilization check on the sensor before completing the calibration. Manual or automatic temperature compensation is menu selectable. Change in pH due to process temper­ ature can be compensated using a programmable tem­ perature coefficient. For more information concerning the use and operation of the pH or ORP sensors, refer to the product data sheets. PERFORMANCE SPECIFICATIONS ­ ANALYZER (ORP INPUT) Model 1056 can also derive an inferred pH value called pHCalc (calculated pH). pHCalc can be derived and displayed when two contacting conductivity sensors are used. (Model 1056­0X­20­30­AN) All standard pH sensors. Measurement Range [ORP]: ­1500 to +1500 mV Accuracy: ± 1 mV Temperature coefficient: ±0.12mV / ºC Input filter: time constant 1 ­ 999 seconds, default 4 seconds. Response time: 5 seconds to 100% of final reading RECOMMENDED SENSORS FOR pH: RECOMMENDED SENSORS FOR ORP: All standard ORP sensors. PERFORMANCE SPECIFICATIONS ­ ANALYZER (pH INPUT) Measurement Range [pH]: 0 to 14 pH Accuracy: ±0.01 pH Diagnostics: glass impedance, reference impedance Temperature coefficient: ±0.002pH/ ºC Solution temperature correction: pure water, dilute base and custom. Buffer recognition: NIST, DIN 19266, JIS 8802, BSI, DIN19267, Ingold, and Merck. Input filter: time constant 1 ­ 999 seconds, default 4 seconds. Response time: 5 seconds to 100% General purpose and high performance pH sensors Models 396PVP, 399VP and 3300HT Temperature Specifications: Temperature range 0­150ºC Temperature Accuracy, Pt­100, 0­50 ºC ± 0.5ºC Temperature Accuracy, Temp. > 50 ºC ± 1ºC 6 MODEL 1056 SECTION 1.0 DESCRIPTION AND SPECIFICATIONS FLOW (Code ­23 and ­33) For use with most pulse signal flow sensors, the 1056 user­selectable units of measurement include flow rates in GPM (Gallons per minute), GPH (Gallon per hour), cu ft/min (cubic feet per min), cu ft/hour (cubic feet per hour), LPM (liters per minute), LPH (liters per hour), or m3/hr (cubic meters per hour), and velocity in ft/sec or m/sec. When configured to measure flow, the unit also acts as a totalizer in the chosen unit (gallons, liters, or cubic meters). Dual flow instruments can be configured as a % recovery, flow difference, flow ratio, or total (combined) flow. Totalized Flow: 0 – 9,999,999,999,999 Gallons or m3, 0 – 999, 999,999,999 cu ft. Accuracy: 0.5% Input filter: time constant 0­999 sec., default 5 sec. RECOMMENDED SENSORS* +GF+ Signet 515 Rotor­X Flow sensor * Input voltage not to exceed ±36V PERFORMANCE SPECIFICATIONS Frequency Range: 3 to 1000 Hz Flow Rate: 0 ­ 99,999 GPM, LPM, m3/hr, GPH, LPH, cu ft/min, cu ft/hr. 4­20mA Current Input (Codes ­23 and ­33) For use with any transmitter or external device that transmits 4­20mA or 0­20mA current outputs. Typical uses are for temperature compensation of live meas­ urements (except ORP, turbidity and flow) and for continuous atmospheric pressure input for determina­ tion of partial pressure, needed for compensation of live dissolved oxygen measurements. External input of atmospheric pressure for DO measurement allows continuous partial pressure compensation while the Model 1056 enclosure is completely sealed. (The pressure transducer component on the DO board can only be used for calibration when the case is open to atmosphere.) Externally sourced current input is also useful for calibration of new or existing sensors that require temperature measurement or atmospheric pressure inputs (DO only). For externally sourced temp or pressure compensation, the user must program the 1056 to input the 4­20mA current signal from the external device. In addition to live continuous compensation of live measurements, the current input board can also be used simply to display the measured temperature. or the calculated partial pressure from the external device. This feature leverages the large display variables on the Model 1056 as a convenience for technicians. Temperature can be displayed in degrees C or degrees F. Partial pressure can be displayed in inches Hg, mm Hg, atm (atmospheres), kPa (kiloPascals), bar or mbar. The current input board can be used with devices that do not actively power their 4­20mA output signals. The Model 1056 actively powers to the + and – lines of the current input board to enable current input from a 4­20mA output device. Note: this Model 1056 signal input board (­23, ­33 model option code) also includes flow measurement functionality. The signal board, however, must be configured to measure either mA current input or flow. PERFORMANCE SPECIFICATIONS Measurement Range *[mA]: 0­20 or 4­20 Accuracy: ±0.03mA Input filter: time constant 0­999 sec., default 5 sec. *Current input not to exceed 22mA 7 MODEL 1056 SECTION 1.0 DESCRIPTION AND SPECIFICATIONS CHLORINE (Code ­24 and ­34) Free and Total Chlorine RECOMMENDED SENSORS The 1056 is compatible with the 499ACL­01 free chlorine sensor and the 499ACL­02 total chlorine sensor. The 499ACL­02 sensor must be used with the TCL total chlorine sample conditioning system. The 1056 fully compensates free and total chlorine readings for changes in membrane permeability caused by temper­ ature changes. For free chlorine measurements, both automatic and manual pH correction are available. For automatic pH correction select code ­32 and an appropri­ ate pH sensor. For more information concerning the use and operation of the amperometric chlorine sensors and the TCL measurement system, refer to the product data sheets. Rosemount Analytical Model 499ACL­03 Monochloramine sensor PERFORMANCE SPECIFICATIONS Resolution: 0.001 ppm or 0.01 ppm – selectable Input Range: 0nA – 100μA Automatic pH correction (requires Code ­32): 6.0 to 10.0 pH Temperature compensation: Automatic (via RTD) or manual (0­50°C). Input filter: time constant 1 ­ 999 sec, default 5 sec. Response time: 6 seconds to 100% of final reading RECOMMENDED SENSORS* Chlorine: Model 499ACL­01 Free Chlorine or Model 499ACL­02 Total Residual Chlorine pH: The following pH sensors are recommended for automatic pH correction of free chlorine readings: Models: 399­09­62, 399­14, and 399VP­09 pH­Independent Free Chlorine The 1056 is compatible with the Model 498CL­01 pH­ independent free chlorine sensor. The Model 498CL­01 sensor is intended for the continuous determination of free chlorine (hypochlorous acid plus hypochlorite ion) in water. The primary application is measuring chlorine in drinking water. The sensor requires no acid pre­treat­ ment, nor is an auxiliary pH sensor required for pH correction. The Model 1056 fully compensates free chlorine readings for changes in membrane permeability caused by temperature. For more information concerning the use and operation of the amperometric chlorine sensors, refer to the product data sheets. PERFORMANCE SPECIFICATIONS Resolution: 0.001 ppm or 0.01 ppm – selectable Input Range: 0nA – 100μA Automatic pH correction: 6.5 to 10.0 pH Temperature compensation: Automatic (via RTD) or manual (0­50°C). Input filter: time constant 1 ­ 999 sec, default 5 sec. Response time: 6 seconds to 100% of final reading RECOMMENDED SENSORS Rosemount Analytical Model 498CL­01 pH independent free chlorine sensor Monochloramine The Model 1056 is compatible with the Model 499A CL­03 Monochloramine sensor. The Model 1056 fully compensates readings for changes in membrane permeability caused by temperature changes. Because monochloramine measurement is not affected by pH of the process, no pH sensor or correction is required. For more information concerning the use and operation of the amperometric chlorine sensors, refer to the product data sheets. PERFORMANCE SPECIFICATIONS Resolution: 0.001 ppm or 0.01 ppm – selectable Input Range: 0nA – 100μA Temperature compensation: Automatic (via RTD) or manual (0­50°C). Input filter: time constant 1 ­ 999 sec, default 5 sec. Response time: 6 seconds to 100% of final reading 8 Chlorine sensors with Variopol connection and cable connection Model 498CL­01 MODEL 1056 SECTION 1.0 DESCRIPTION AND SPECIFICATIONS DISSOLVED OXYGEN (Codes ­25 and ­35) DISSOLVED OZONE (Code ­26 and ­36) The 1056 is compatible with the 499ADO, 499ATrDO, Hx438, and Gx438 dissolved oxygen sensors and the 4000 percent oxygen gas sensor. The 1056 displays dissolved oxygen in ppm, mg/L, ppb, μg/L, % satura­ tion, % O2 in gas, ppm O2 in gas. The analyzer fully compensates oxygen readings for changes in mem­ brane permeability caused by temperature changes. An atmospheric pressure sensor is included on all dis­ solved oxygen signal boards to allow automatic atmos­ pheric pressure determination at the time of calibration. If removing the sensor from the process liquid is imprac­ tical, the analyzer can be calibrated against a standard instrument. Calibration can be corrected for process salinity. For more information on the use of ampero­ metric oxygen sensors, refer to the product data sheets. The 1056 is compatible with the Model 499AOZ sen­ sor. The 1056 fully compensates ozone readings for changes in membrane permeability caused by temper­ ature changes. For more information concerning the use and operation of the amperometric ozone sensors, refer to the product data sheets. PERFORMANCE SPECIFICATIONS RECOMMENDED SENSOR Resolution: 0.01 ppm; 0.1 ppb for 499A TrDO sensor (when O2 <1.00 ppm); 0.1% Input Range: 0nA – 100μA Temperature Compensation: Automatic (via RTD) or manual (0­50°C). Input filter: time constant 1 ­ 999 sec, default 5 sec. Response time: 6 seconds to 100% of final reading Rosemount Analytical Model 499A OZ ozone sensor PERFORMANCE SPECIFICATIONS Resolution: 0.001 ppm or 0.01 ppm – selectable Input Range: 0nA – 100μA Temperature Compensation: Automatic (via RTD) or manual (0­35°C) Input filter: time constant 1 ­ 999 sec, default 5 sec. Response time: 6 seconds to 100% of final reading RECOMMENDED SENSORS Rosemount Analytical amperometric membrane and steam­sterilizable sensors listed above Dissolved Ozone sensors with Polysulfone body Variopol connection and cable connection Model 499AOZ Dissolved Oxygen sensor with Variopol connection Model 499ADO 9 MODEL 1056 SECTION 1.0 DESCRIPTION AND SPECIFICATIONS Turbidity (Codes ­27 and ­37) The 1056 instrument is available in single and dual tur­ bidity configurations for the Clarity II® turbidimeter. It is intended for the determination of turbidity in filtered drinking water. The other components of the Clarity II turbidimeter – sensor(s), debubbler/measuring cham­ ber(s), and cable for each sensor must be ordered separately or as a complete system with the Model 1056. The 1056 turbidity instrument accepts inputs from both USEPA 180.1 and ISO 7027­compliant sensors When ordering the Model 1056 turbidity instrument, the ­02 (24VDC power supply) or the ­03 (switching 115/230VAC power supply) are required. Both of these power supplies include four fully programmable relays with timers. Note: Model 1056 Turbidity must be used with Clarity II sensor, sensor cable and debubbler. PERFORMANCE SPECIFICATIONS Units: Turbidity (NTU, FTU, or FNU); total suspended solids (mg/L, ppm, or no units) Display resolution­turbidity: 4 digits; decimal point moves from x.xxx to xxx.x Display resolution­TSS: 4 digits; decimal point moves from x.xxx to xxxx Calibration methods: user­prepared standard, com­ mercially prepared standard, or grab sample. For total suspended solids user must provide a linear calibration equation. Inputs: Choice of single or dual input, EPA 180.1 or ISO 7027 sensors. Field wiring terminals: removable terminal blocks for sensor connection. Accuracy after calibration at 20.0 NTU: 0­1 NTU ±2% of reading or 0.015 NTU, whichever is greater. 0­20 NTU: ±2% of reading. Clarity ll Turbidimeter 10 MODEL 1056 SECTION 2.0 INSTALLATION SECTION 2.0. INSTALLATION 2.1 UNPACKING AND INSPECTION 2.2 INSTALLATION 2.1 UNPACKING AND INSPECTION Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions. Save the box. If there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present. If items are missing, notify Rosemount Analytical immediately. 2.2 INSTALLATION 2.2.1 General Information 1. Although the analyzer is suitable for outdoor use, do not install it in direct sunlight or in areas of extreme tem­ peratures. 2. Install the analyzer in an area where vibration and electromagnetic and radio frequency interference are min­ imized or absent. 3. Keep the analyzer and sensor wiring at least one foot from high voltage conductors. Be sure there is easy access to the analyzer. 4. The analyzer is suitable for panel, pipe, or surface mounting. Refer to the table below. Type of Mounting Figure Panel 2­1 Wall and Pipe 2­2 WARNING RISK OF ELECTRICAL SHOCK Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA­70) and/or any other applicable national or local codes. 11 FIGURE 2­1 PANEL MOUNTING DIMENSIONS MILLIMETER INCH 17.13 1.1 126.4 5.0 101.6 4.00 154.9 6.1 154.9 6.1 Front View Side View ( 126.4 5.0 ) 76.2 3.0 41.4 1.6 Bottom View 152.73 6.0 Note: Panel mounting seal integrity (4/4X) for outdoor applications is the responsibility of the end user. 12 FIGURE 2­2 PIPE AND WALL MOUNTING DIMENSIONS (Mounting bracket PN:23820­00) MILLIMETER INCH 154.9 6.1 Wall / Surface Mount 232 9.1 102 4.0 33.5 1.3 130 5.1 187 7.4 154.9 6.1 165 6.5 Side View Front View Pipe Mount 232 9.1 Bottom View 33.5 1.3 130 5.1 80.01 3.2 165 6.5 45.21 1.8 108.9 4.3 Side View 71.37 2.8 The front panel is hinged at the bottom. The panel swings down for easy access to the wiring locations. 13 FIGURE 2­3 CSA Non­Incendive Class I, Division 2 Certified product for selected configurations (for approved models, see Fig. 2­4) MODEL 1056 14 SECTION 2.0 INSTALLATION FIGURE 2­4 CSA Non­Incendive Class I, Division 2 Certified product for selected configurations MODEL 1056 SECTION 2.0 INSTALLATION 15 FIGURE 2­5 FM Non­Incendive Class I, Division 2 Certified product for selected configurations (for approved models, see Fig. 2­6) MODEL 1056 16 SECTION 2.0 INSTALLATION FIGURE 2­6 FM Non­Incendive Class I, Division 2 Certified product for selected configurations MODEL 1056 SECTION 2.0 INSTALLATION 17 MODEL 1056 18 SECTION 2.0 INSTALLATION This page left blank intentionally MODEL 1056 SECTION 3.0 WIRING SECTION 3.0. WIRING 3.1 3.2 3.3 3.4 GENERAL PREPARING CONDUIT OPENINGS PREPARING SENSOR CABLE POWER, OUTPUT, AND SENSOR CONNECTIONS 3.1 GENERAL The 1056 is easy to wire. It includes removable connectors and slide­out signal input boards. The front panel is hinged at the bottom. The panel swings down for easy access to the wiring locations. 3.1.1. Removable connectors and signal input boards Model 1056 uses removable signal input boards and communication boards for ease of wiring and instal­ lation. Each of the signal input boards can be partially or completely removed from the enclosure for wiring. The Model 1056 has three slots for placement of up to two signal input boards and one communication board. Slot 1­Left Slot 2 – Center Slot 3 – Right Comm. board Input Board 1 Input Board 2 3.1.2. Signal Input boards Slots 2 and 3 are for signal input measurement boards. Wire the sensor leads to the measurement board following the lead locations marked on the board. After wiring the sensor leads to the signal board, carefully slide the wired board fully into the enclosure slot and take up the excess sensor cable through the cable gland. Tighten the cable gland nut to secure the cable and ensure a sealed enclosure. 3.1.3. Digital Communication boards HART and Profibus DP communication boards will be available in the future as options for Model 1056 digital communication with a host. The HART board supports Bell 202 digital communications over an analog 4­20mA current output. Profibus DP is an open communications protocol which operates over a dedicated digital line to the host. 3.1.4 Alarm relays Four alarm relays are supplied with the switching power supply (85 to 265VAC, ­03 order code) and the 24VDC power supply (20­30VDC, ­02 order code). All relays can be used for process measurement(s) or temperature. Any relay can be configured as a fault alarm instead of a process alarm. Each relay can be configured independently and each can be programmed as an interval timer, typically used to activate pumps or control valves. As process alarms, alarm logic (high or low activation or USP*) and deadband are user­programmable. Customer­defined failsafe operation is supported as a programmable menu function to allow all relays to be energized or not­energized as a default condition upon powering the analyzer. The USP* alarm can be programmed to activate when the conductivity is within a user­selectable percentage of the limit. USP alarming is available only when a contacting conductivity measurement board is installed. 3.2 PREPARING CONDUIT OPENINGS There are six conduit openings in all configurations of Model 1056. (Note that four of the openings will be fitted with plugs upon shipment.) Conduit openings accept 1/2­inch conduit fittings or PG13.5 cable glands. To keep the case watertight, block unused openings with NEMA 4X or IP65 conduit plugs. NOTE: Use watertight fittings and hubs that comply with your requirements. Connect the conduit hub to the conduit before attaching the fitting to the analyzer. 19 MODEL 1056 SECTION 3.0 WIRING 3.3 PREPARING SENSOR CABLE The 1056 is intended for use with all Rosemount Analytical sensors. Refer to the sensor installation instructions for details on preparing sensor cables. 3.4 POWER, OUTPUT, AND SENSOR CONNECTIONS 3.4.1 Power wiring Three Power Supplies are offered for Model 1056: a. 115/230VAC Power Supply (­01 ordering code) b. 24VDC (20 – 30V) Power Supply (­02 ordering code) c. 85 – 265 VAC Switching Power Supply (­03 ordering code) AC mains (115 or 230V) leads and 24VDC leads are wired to the Power Supply board which is mounted vertically on the left side of the main enclosure cavity. Each lead location is clearly marked on the Power Supply board. Wire the power leads to the Power Supply board using the lead markings on the board. The grounding plate is connected to the earth terminal of power supply input connector TB1 on the ­01 (115/230VAC) and ­03 (85­265VAC) power supplies. The green colored screws on the grounding plate are intend­ ed for connection to some sensors to minimize radio frequency interference. The green screws are not intended to be used for safety purposes. 115/230VAC Power Supply (­01 ordering code) is shown below: CAUTION AC Power switch shipped in the 230VAC position. Adjust switch upwards to 115VAC position for 110VAC – 120VAC operation. Figure 3­1 24VDC Power Supply (­02 ordering code) is shown below: This power supply automatically detects DC power and accepts 20VDC to 30VDC inputs. Four programmable alarm relays are included. Figure 3­2 Switching AC Power Supply (­03 ordering code) is shown below: This power supply automatically detects AC line condi­ tions and switches to the proper line voltage and line frequency. Four programmable alarm relays are included. 20 Figure 3­3 MODEL 1056 SECTION 3.0 WIRING 3.4.2 Current Output wiring All instruments are shipped with two 4­20mA current outputs. Wiring locations for the outputs are on the Main board which is mounted on the hinged door of the instrument. Wire the out put leads to the correct posi­ tion on the Main board using the lead markings (+/positive, ­/negative) on the board. Male mating connectors are provided with each unit. Figure 3.4 3.4.3 Alarm relay wiring Four alarm relays are supplied with the switching power supply (85 to 265VAC, ­03 order code) and the 24VDC power supply (20­30VDC, ­02 order code). Wire the relay leads on each of the independent relays to the correct position on the power supply board using the printed lead markings (NO/Normally Open, NC/Normally Closed, or Com/Common) on the board. See Fig 3­4. NO1 COM1 RELAY 1 NC1 NO2 COM2 RELAY 2 NC2 NO3 COM3 RELAY 3 NC3 NO4 COM4 RELAY 4 NC4 Figure 3­5 Alarm Relay Wiring for Model 1056 Switching Power Supply (­03 Order Code) 3.4.4 Sensor wiring to signal boards Wire the correct sensor leads to the measurement board using the lead locations marked directly on the b o a r d . After wiring the sensor leads to the signal board, carefully slide the wired board fully into the enclosure slot and take up the excess sensor cable through the cable gland. For best EMI/RFI protection use shielded output signal cable enclosed in an earth­grounded metal conduit. Connect the shield to earth ground. AC wiring should be 14 gauge or greater. Provide a switch or breaker to dis­ connect the analyzer from the main power supply. Install the switch or breaker near the analyzer and label it as the disconnecting device for the analyzer. Keep sensor and output signal wiring separate from power wiring. Do not run sensor and power wiring in the same conduit or close together in a cable tray. WARNING RISK OF ELECTRICAL SHOCK Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA­70) and/or any other applicable national or local codes. 21 MODEL 1056 SECTION 3.0 WIRING Sec. 3.4 Signal board wiring Figure 3­6 Contacting Conductivity signal board and Sensor cable leads Figure 3­7 Toroidal Conductivity Signal board and Sensor cable leads 22 MODEL 1056 SECTION 3.0 WIRING Figure 3­8 pH/ORP/ISE signal board and Sensor cable leads Figure 3­9 Amperometric signal (Chlorine, Oxygen, Ozone) board and Sensor cable leads 23 MODEL 1056 SECTION 3.0 WIRING Figure 3­10 Turbidity signal board with plug­in Sensor connection Figure 3­11 Flow/Current Input signal board and Sensor cable leads 24 MODEL 1056 SECTION 3.0 WIRING FIGURE 3­12 Power Wiring for the 1056 115/230VAC Power Supply (­01 Order Code) FIGURE 3­13 Power Wiring for the 1056 85­265 VAC Power Supply (­03 ordering code) 25 MODEL 1056 SECTION 3.0 WIRING FIGURE 3­14 Output Wiring for Model 1056 Main PCB To Main PCB FIGURE 3­15 Power Wiring for Model 1056 24VDC Power Supply (­02 ordering code) 26 MODEL 1056 SECTION 4.0 DISPLAY AND OPERATION SECTION 4.0 DISPLAY AND OPERATION 4.1 4.2 4.3 4.4 USER INTERFACE KEYPAD MAIN DISPLAY MENU SYSTEM 4.1 USER INTERFACE The 1056 has a large display which shows two live measurement readouts in large digits and up to four additional process variables or diagnostic parameters concurrently. The display is back­lit and the format can be customized to meet user requirements. The intu­ itive menu system allows access to Calibration, Hold (of current outputs), Programming, and Display functions by pressing the MENU button. In addition, a dedicated DIAGNOSTIC button is available to provide access to useful operational information on installed sensor(s) and any problematic conditions that might occur. The display flashes Fault and/or Warning when these condi­ tions occur. Help screens are displayed for most fault and warning conditions to guide the user in trou­ bleshooting. During calibration and programming, key presses cause different displays to appear. The displays are self­ explanatory and guide the user step­by­step through the procedure. 4.2 INSTRUMENT KEYPAD There are 4 Function keys and 4 Selection keys on the instrument keypad. Function keys: The MENU key is used to access menus for program­ ming and calibrating the instrument. Four top­level menu items appear when pressing the MENU key:  Calibrate: calibrate attached sensors and analog outputs.  Hold: Suspend current outputs.  Program: Program outputs, measurement, temperature, security and reset.  Display: Program display format, language, warnings, and contrast Pressing MENU always causes the main menu screen to appear. Pressing MENU followed by EXIT causes the main display to appear. 27 MODEL 1056 Pressing the DIAG key displays active Faults and Warnings, and provides detailed instrument information and sensor diagnostics including: Faults, Warnings, Sensor 1 and 2 information, Out 1 and Out 2 live current values, model configuration string e.g. 1056­01­20­31­ AN, Instrument Software version, and AC frequency used. Pressing ENTER on Sensor 1 or Sensor 2 pro­ vides useful diagnostics and information (as applica­ ble): Measurement, Sensor Type, Raw signal value, Cell constant, Zero Offset, Temperature, Temperature SECTION 4.0 DISPLAY AND OPERATION Offset, selected measurement range, Cable Resistance, Temperature Sensor Resistance, Signal Board software version. The ENTER key. Pressing ENTER stores numbers and settings and moves the display to the next screen. The EXIT key. Pressing EXIT returns to the previous screen without storing changes. Selection keys: Surrounding the ENTER key, four Selection keys – up, down, right and left, move the cursor to all areas of the screen while using the menus. Selection keys are used to: 1. select items on the menu screens 2. scroll up and down the menu lists. 3. enter or edit numeric values. 4. move the cursor to the right or left 5. select measurement units during operations 4.3 MAIN DISPLAY The Model 1056 displays one or two primary measurement values, up to four secondary measurement values, a fault and warning banner, alarm relay flags, and a digital communications icon. Process measurements: Two process variables are displayed if two signal boards are installed. One process variable and process tem­ perature is displayed if one signal board is installed with one sensor. The Upper display area shows the Sensor 1 process reading. The Center display area shows the Sensor 2 process reading. For dual conductivity, the Upper and Center display areas can be assigned to different process variables as follows: Process variables for Upper display­ example: Process variables for Center display­ example: Measure 1 Measure 1 % Reject Measure 2 % Pass % Reject Ratio % Pass Ratio Blank For single input configurations, the Upper display area shows the live process variable and the Center display area can be assigned to Temperature or blank. Secondary values: Up to four secondary values are shown in four display quadrants at the bottom half of the screen. All four secondary value positions can be programmed by the user to any display parameter available. Possible secondary values include: Displayable Secondary Values Slope 1 Man Temp 2 Ref Off 1 Output 1 mA Gl Imp 1 Output 2 mA Ref Imp 1 Output 1 % Raw Output 2 % mV Input Measure 1 Temp 1 Blank Man Temp 1 28 MODEL 1056 SECTION 4.0 DISPLAY AND OPERATION Fault and Warning banner: If the analyzer detects a problem with itself or the sensor the word Fault or Warning will appear at the bottom of the display. A fault requires immediate attention. A warning indicates a problematic condition or an impending fail­ ure. For troubleshooting assitance, press Diag. Formatting the Main Display The main display screen can be programmed to show primary process variables, secondary process variables and diagnostics. 1. Press MENU 2. Scroll down to Display. Press ENTER. 3. Main Format will be highlighted. Press ENTER. 4. The sensor 1 process value will be highlighted in reverse video. Press the selection keys to navigate down to the screen sections that you wish to program. Press ENTER. 5. Choose the desired display parameter or diagnostic for each of the four display sections in the lower screen. 6. Continue to navigate and program all desired screen sections. return to the main display. Press MENU and EXIT. The screen will For single sensor configurations, the default display shows the live process measurement in the upper display area and temperature in the center display area. The user can elect to disable the display of temperature in the cen­ ter display area using the Main Format function. See Fig. 4­1 to guide you through programming the main display to select process parameters and diagnostics of your choice. For dual sensor configurations, the default display shows Sensor 1 live process measurement in the upper display area and Sensor 2 live process measurement temperature in the center display area. See Fig. 4­1 to guide you through programming the main display to select process parameters and diagnostics of your choice. 4.4 MENU SYSTEM Model 1056 uses a scroll and select menu system. Pressing the MENU key at any time opens the top­level menu including Calibrate, Hold, Program and Display functions. To find a menu item, scroll with the up and down keys until the item is highlighted. Continue to scroll and select menu items until the desired function is chosen. To select the item, press ENTER. To return to a previ­ ous menu level or to enable the main live display, press the EXIT key repeatedly. To return immediately to the main display from any menu level, simply press MENU then EXIT. The selection keys have the following functions:  The Up key (above ENTER) increments numerical values, moves the decimal place one place to the right, or selects units of measurement.  The Down key (below ENTER) decrements numerical values, moves the decimal place one place to the left, or selects units of measurement   The Left key (left of ENTER) moves the cursor to the left. The Right key (right of ENTER) moves the cursor to the right. To access desired menu functions, use the “Quick Reference” Figure B. During all menu displays (except main display format and Quick Start), the live process measurements and secondary measurement values are displayed in the top two lines of the Upper display area. This conveniently allows display of the live values during important calibration and programming operations. Menu screens will time out after two minutes and return to the main live display. 29 MODEL 1056 SECTION 4.0 DISPLAY AND OPERATION FIGURE 4­1 Formatting the Main Display 30 MODEL 1056 SECTION 5.0 PROGRAMMING THE ANALYZER ­ BASICS SECTION 5.0. PROGRAMMING THE ANALYZER ­ BASICS 5.1 GENERAL 5.2 CHANGING START­UP SETTINGS 5.3 PROGRAMMING TEMPERATURE 5.4 CONFIGURING AND RANGING 4­20MA OUTPUTS 5.5 SETTING SECURITY CODES 5.6 SECURITY ACCESS 5.7 USING HOLD 5.8 RESETTING FACTORY DEFAULTS – RESET ANALYZER 5.9 PROGRAMMING ALARM RELAYS 5.1 GENERAL Section 5.0 describes the following programming functions:  Changing the measurement type, measurement units and temperature units.  Choose temperature units and manual or automatic temperature compensation mode  Configure and assign values to the current outputs  Set a security code for two levels of security access  Accessing menu functions using a security code  Enabling and disabling Hold mode for current outputs  Choosing the frequency of the AC power (needed for optimum noise rejection)  Resetting all factory defaults, calibration data only, or current output settings only 5.2 CHANGING STARTUP SETTINGS 5.2.1 Purpose To change the measurement type, measurement units, or temperature units that were initially entered in Quick Start, choose the Reset analyzer function (Sec. 5.9) or access the Program menus for sensor 1 or sensor 2 (Sec. 6.0). The following choices for specific measurement type, measurement units are available for each sensor meas­ urement board. TABLE 5­1. Measurements and Measurement Units Signal board pH/ORP (­22, ­32) Contacting conductivity (­20, ­30) Toroidal conductivity (­21, ­31) Available measurements Measurements units: pH, ORP, Redox, Ammonia, Fluoride, Custom ISE Conductivity, Resistivity, TDS, Salinity, NaOH (0­12%), HCl (0­15%), Low H2SO4, High H2SO4, NaCl (0­20%), Custom Curve pH, mV (ORP) %, ppm, mg/L, ppb, μg/L, (ISE) Conductivity, Resistivity, TDS, Salinity, NaOH (0­12%), HCl (0­15%), Low H2SO4, High H2SO4, NaCl (0­20%), Custom Curve Chlorine (­24, ­34) Oxygen (­25, ­35) Free Chlorine, pH Independ. Free Cl, Total Chlorine, Monochloramine Ozone (­26, ­36) Temperature (all) Ozone Temperature Oxygen (ppm), Trace Oxygen (ppb), Percent Oxygen in gas, Salinity μS/cm, mS/cm, S/cm % (concentration) μS/cm, mS/cm, S/cm % (concentration) ppm, mg/L ppm, mg/L, ppb, µg/L % Sat, Partial Pressure, % Oxygen In Gas, ppm Oxygen In Gas ppm, mg/L, ppb, μg/L °C. ºF 5.2.2 Procedure. Follow the Reset Analyzer procedure (Sec 5.8) to reconfigure the analyzer to display new measurements or measurement units. To change the specific measurement or measurement units for each signal board type, refer to the Program menu for the appropriate measurement (Sec. 6.0). 31 MODEL 1056 SECTION 5.0 PROGRAMMING THE ANALYZER ­ BASICS 5.3 CHOOSING TEMPERATURE UNITS AND AUTOMATIC/MANUAL TEMPERATURE COMPENSATION 5.3.1 Purpose Most liquid analytical measurements (except ORP) require temperature compensation. The Model 1056 performs temperature compensation automatically by applying internal temperature correction algorithms. Temperature correction can also be turned off. If tem­ perature correction is off, the Model 1056 uses the tem­ perature entered by the user in all temperature correc­ tion calculations. 5.3.2 Procedure. Follow the menu screens in Fig. 5.1 to select automatic or manual temp compensation, set the manual reference temperature, and to program temperature units as °C or °F. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC Temperature Units: °C S1 Temp Comp: Auto S2 Temp Comp: Auto S1 Manual: +25.0°C S2 Manual: +25.0ºC Figure 5­1. Choosing Temp Units and Manual Auto Temp Compensation 5.4 CONFIGURING AND RANGING THE CURRENT OUTPUTS 5.4.1 Purpose The Model 1056 accepts inputs from two sensors and has two analog current outputs. Ranging the outputs means assigning values to the low (0 or 4 mA) and high (20 mA) outputs. This section provides a guide for configuring and ranging the outputs. ALWAYS CONFIGURE THE OUTPUTS FIRST. 5.4.2 Definitions 1. CURRENT OUTPUTS. The analyzer provides a con­ tinuous output current (4­20 mA or 0­20 mA) directly proportional to the process variable or temperature. 32 The low and high current outputs can be set to any value. 2. ASSIGNING OUTPUTS. Assign a measurement to Output 1 or Output 2. 3. DAMPEN. Output dampening smooths out noisy readings. It also increases the response time of the output. Output dampening does not affect the response time of the display. 4. MODE. The current output can be made directly proportional to the displayed value (linear mode) or directly proportional to the common logarithm of the displayed value (log mode). MODEL 1056 SECTION 5.0 PROGRAMMING THE ANALYZER ­ BASICS 5.4.3 Procedure: Configure Outputs. Under the Program/Outputs menu, the adjacent screen will appear to allow configuration of the outputs. Follow the menu screens in Fig. 5­2 to configure the outputs. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC OutputM Configure Assign: S1 Meas Range: 4­20mA Scale: Linear Dampening: 0sec Fault Mode: Fixed Fault Value: 21.00mA 5.4.4 Procedure: Assigning Measurements the Low and High Current Outputs The adjacent screen will appear when entering the Assign function under Program/Output/Configure. These screens allow you to assign a measurement, process value, or temperature input to each output. Follow the menu screens in Fig. 5­2 to assign measurements to the outputs. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC 5.4.5 Procedure: Ranging the Current Outputs The adjacent screen will appear under Program/Output/Range. Enter a value for 4mA and 20mA (or 0mA and 20mA) for each output. Follow the menu screens in Fig. 5­2 to assign values to the out­ puts. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC OutputM Assign S1 Measurement S1 Temperature S2 Measurement S2 Temperature OM OM OM OM SN SN SN SN Output Range 4mA: 0.000µS/cm 20mA: 20.00µS/cm 4mA: 00.00pH 20mA: 14.00pH Figure 5­2. Configuring and Ranging the Current Outputs 33 MODEL 1056 SECTION 5.0 PROGRAMMING THE ANALYZER ­ BASICS 2. 3. 5.5 SETTING A SECURITY CODE 5.5.1 Purpose. The security codes prevent accidental or unwanted changes to program settings, displays, and calibration. Model 1056 has two levels of security code to control access and use of the instrument to different types of users. The two levels of security are: ­ All: This is the Supervisory security level. It allows access to all menu functions, including Programming, Calibration, Hold and Display. ­ Calibration/Hold: This is the operator or technician level menu. It allows access to only calibration and Hold of the current outputs. 5.5.2 Procedure. 1. Press MENU. The main menu screen appears. Choose Program. Scroll down to Security. Select Security. The security entry screen appears. Enter a three digit security code for each of the desired security levels. The security code takes effect two minutes after the last key stroke. Record the security code(s) for future access and communication to operators or techni­ cians as needed. 4. The display returns to the security menu screen. Press EXIT to return to the previous screen. To return to the main display, press MENU followed by EXIT. Fig. 5­3 displays the security code screens. Program MAIN MENU Figure 5­3. Setting a Security Code S1: 1.234µS/cm S2: 12.34pH 25.0ºC 25.0ºC Program Outputs Measurement Temperature S1: 1.234µS/cm S2: 12.34pH Security Diagnostic Setup Ambient AC Power:Unk Reset Analyzer 34 25.0ºC 25.0ºC Security Calibration/Hold: 000 All: 000 MODEL 1056 SECTION 5.0 PROGRAMMING THE ANALYZER ­ BASICS 5.6 SECURITY ACCESS 5.7 USING HOLD 5.6.1 How the Security Code Works When entering the correct access code for the Calibration/Hold security level, the Calibration and Hold menus are accessible. This allows operators or technicians to perform routine maintenance. This security level does not allow access to the Program or Display menus. When entering the correct access code for All security level, the user has access to all menu functions, includ­ ing Programming, Calibration, Hold and Display. 5.7.1 Purpose The analyzer output is always proportional to measured value. To prevent improper operation of systems or pumps that are controlled directly by the current output, place the analyzer in hold before removing the sensor for calibration and maintenance. Be sure to remove the analyzer from hold once calibration is complete. During hold, both outputs remain at the last value. Once in hold, all current outputs remain on Hold indefinitely. 5.6.2 Procedure. 1. If a security code has been programmed, selecting the Calibrate, Hold, Program or Display top menu items causes the security access screen to appear 5.7.2 Using the Hold Function To hold the outputs, 1. Press MENU. The main menu screen appears. Choose Hold. 2. The Hold Outputs and Alarms? screen appears. Choose Yes to place the analyzer in hold. Choose No to take the analyzer out of hold. Note: There are no alarm relays with this con figuration. Current outputs are included with all configurations. 3. The Hold screen will then appear and Hold will remain on indefinitely until Hold is disabled. 2. Enter the three­digit security code for the appropriate security level. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC Security Code 000 3. If the entry is correct, the appropriate menu screen appears. If the entry is incorrect, the Invalid Code screen appears. The Enter Security Code screen reappears after 2 seconds. See figure 5­1 below. Hold MAIN MENU Figure 5­4. Using Hold S1: 1.234µS/cm S2: 12.34pH 25.0ºC 25.0ºC Hold S1 Hold: S2 Hold: No No S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC S1 Hold outputs and alarms? No Yes 35 MODEL 1056 SECTION 5.0 PROGRAMMING THE ANALYZER ­ BASICS 5.8 RESETTING FACTORY DEFAULT SETTINGS 5.8.1 Purpose. This section describes how to restore factory calibration and default values. The process also clears all fault messages and returns the display to the first Quick Start screen. The Model 1056 offers three options for resetting factory defaults. a. reset all settings to factory defaults b. reset sensor calibration data only c. reset analog output settings only 5.8.2. Procedure. To reset to factory defaults, reset calibration data only or reset analog outputs only, follow the Reset Analyzer flow diagram. Figure 5­5. Resetting Factory Default Settings 36 MODEL 1056 SECTION 5.0 PROGRAMMING THE ANALYZER ­ BASICS 5.9 Programming Alarm Relays 5.9.1 Purpose. The Model 1056 24VDC (­02 order code) and the AC switching power supply (­03 order code) provide four alarm relays for process measurement or temperature. Each alarm can be configured as a fault alarm instead of a process alarm. Also, each relay can be programmed independently and each can be programmed as an interval timer. This section describes how to configure alarm relays, simulate relay activation, and synchronize timers for the four alarm relays. This section provides details to program the following alarm features: Sec. Alarm relay feature: default Description 5.9.2 Enter Setpoint 100.0uS/cm Enter alarm trigger value 5.9.3 Assign measurement S1 Measure Select alarm assignment 5.9.4 Set relay logic Program relay to activate at High or Low reading High 0.00uS/cm Program the change in process value after the relay deactivates 5.9.5 Deadband: 5.9.6 USP Safety: 5.9.7 Normal state: 5.9.8 Interval time: 24.0 hr Time in hours between relay activations 5.9.9 On­Time: 10 min Enter the time in seconds that the relay is activated. 5.9.10 Recover time: 60 sec Enter time after the relay deactivation for process recovery 5.9.11 Hold while active: 5.9.12 Simulate 5.9.13 Synchronize Timers 0%↓ Open S1 Program percentage of the limit to activate the alarm Program relay default condition as open or closed for failsafe operation Holds current outputs during relay activation Manually simulate alarms to confirm relay operation Yes Control the timing of two or more relay timers set as Interval timers Under the Program/Alarms menu, this screen will appear to allow configuration of the alarm relays. Follow the menu screens in Fig. XX to configure the outputs. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC This screen will appear to allow selection of a specific alarm relay. Select the desired alarm and press ENTER. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC This screen will appear next to allow complete pro­ gramming of each alarm. Factory defaults are dis­ played as they would appear for an installed contact­ ing conductivity board. USP Safety only appears if alarm logic is set to “USP”. Interval timer, On Time, Recover Time, and Hold While Active only appear if the alarm is configured as an Interval timer. Alarms Configure/Setpoint Simulate Synchronize Timers: Yes Configure/Setpoint Alarm 1 Alarm 2 Alarm 3 Alarm 4 S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC AlarmM Settings Setpoint: 100.0uS/cm Assign: S1 Measure Logic: High Deadband: 0.00uS/cm USP Safety: 0%↓ Interval time: 24.0 hr On Time: 120 sec Recover time: 60 sec Hold while active: Sens1 37 MODEL 1056 5.9.2 Procedure – Enter Setpoints Under the Program/Alarms menu, this screen will appear to allow configuration of the alarm relays. Enter the desired value for the process measurement or temperature at which to activate an alarm event. 5.9.3 Procedure – Assign Measurement Under the Alarms Settings menu, this screen will appear to allow assignment of the alarm relays. select an alarm assignment. Additional assignment choices are shown in Figure X­X depending on which meas­ urement board(s) is installed. SECTION 5.0 PROGRAMMING THE ANALYZER ­ BASICS S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC Alarm1 S2 Setpoint +100.0uS/cm S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC AlarmM Assign: S1 Measurement S1 Temperature S2 Measurement S2 Temperature Interval Timer Fault Off 5.9.4 Procedure – Set Relay Logic Under the Alarms Settings menu, this screen will appear to set the alarm logic. Select the desired relay logic to activate alarms at a High reading or a Low reading. USP Safety only appears if a contacting con­ ductivity board is installed. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC 5.9.5 Procedure – Deadband Under the Alarms Settings menu, this screen will appear to program the deadband as a measurement value. Enter the change in the process value needed after the relay deactivates to return to normal (and thereby preventing repeated alarm activation). S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC 5.9.6 Procedure – USP Safety Under the Alarms Settings menu, this screen will appear to program the USP alarm setting. Enter the percentage below the limit at which to activate the alarm. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC 5.9.7 Procedure – Normal state The user can define failsafe condition in software by programming the alarm default state to normally open or normally closed upon power up. To display this alarm configuration item, enter the Expert menus by holding down the EXIT key for 6 seconds while in the main display mode. Select Yes upon seeing the screen prompt: “Enable Expert Menu?” Under the Alarms Settings menu, this screen will appear to set the normal state of the alarms. Select the alarm condition that is desired each time the analyzer is powering up. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC 38 AlarmM Logic: High Low USP Alarm1 Deadband +000.5uS/cm Alarm1 USP Safety +0%i Alarm2 Normal State Open Closed MODEL 1056 SECTION 5.0 PROGRAMMING THE ANALYZER ­ BASICS 5.9.8 Procedure – Interval time Under the Alarms Settings menu, this screen will appear to set the interval time. Enter the fixed time in hours between relay activations. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC 5.9.9 Procedure – On time Under the Alarms Settings menu, this screen will appear to set the relay on time. Enter the time in sec­ onds that the relay is activated. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC 5.9.10 Procedure – Recovery time Under the Alarms Settings menu, this screen will appear to set the relay recovery time. Enter time after the relay deactivation for process recovery. Alarm1 Interval Time 024.0 hrs Alarm1 On­Time 00.00sec S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC Alarm1 Recovery 060sec 5.9.11 Procedure – Hold while active Under the Alarms Settings menu, this screen will appear to program the feature that Holds the current outputs while alarms are active. Select to hold the current outputs for Sensor 1, Sensor 2 or both sensors while the relay is activated. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC 5.9.12 Procedure – Simulate Alarm relays can be manually set for the purposes of checking devices such as valves or pumps. Under the Alarms Settings menu, this screen will appear to allow manual forced activation of the alarm relays. Select the desired alarm condition to simulate. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC 5.9.13 Procedure – Synchronize Under the Alarms Settings menu, this screen will appear to allow Synchronization of alarms that are set to interval timers. Select yes or no to Synchronize two or more timers. Alarm1 Hold while active Sensor 1 Sensor 2 Both None Simulate Alarm M Don’t simulate De­energize Energize S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC Synchronize Timers Yes No 39 MODEL 1056 40 SECTION 5.0 PROGRAMMING THE ANALYZER ­ BASICS This page left blank intentionally MODEL 1056 SECTION 6.0 PROGRAMMING THE MEASUREMENTS SECTION 6.0 PROGRAMMING ­ MEASUREMENTS 6.1 6.2 6.3 6.4 6.5 6.6 CONFIGURING MEASUREMENTS – INTRODUCTION pH ORP CONTACTING CONDUCTIVITY TOROIDAL CONDUCTIVITY CHLORINE 6.6.1 FREE CHLORINE 6.6.2 TOTAL CHLORINE 6.6.3 MONOCHLORAMINE 6.6.4 pH­INDEPENDENT FREE CHLORINE 6.7 OXYGEN 6.8 OZONE 6.9 TURBIDITY 6.10 FLOW 6.11 CURRENT INPUT 6.1 PROGRAMMING MEASUREMENTS – INTRODUCTION The Model 1056 automatically recognizes each installed measurement board upon first power­up and each time the analyzer is powered. Completion of Quick Start screens upon first power up enable measurements, but addi­ tional steps may be required to program the analyzer for the desired measurement application. This section cov­ ers the following programming and configuration functions; 1. Selecting measurement type or sensor type (all sections) 2. Identifying the preamp location (pH­see Sec. 6.2) 3. Enabling manual temperature correction and entering a reference temperature (all sections) 4. Enabling sample temperature correction and entering temperature correction slope (selected sections) 5. Defining measurement display resolution (pH and amperometric) 6. Defining measurement display units (all sections) 7. Adjusting the input filter to control display and output reading variability or noise (all sections) 8. Selecting a measurement range (conductivity – see Sec’s 6.4, 6.5) 9. Entering a cell constant for a contacting or toroidal sensor (see Sec’s 6.4, 6.5) 10. Entering a temperature element/RTD offset or temperature slope (conductivity­see Sec’s 6.4) 11. Creating an application­specific concentration curve (conductivity­see Sec’s 6.4, 6.5) 12. Enabling automatic pH correction for free chlorine measurement (Sec. 6.6.1) To fully configure the analyzer for each installed measurement board, you may use the following: 1. Reset Analyzer function to reset factory defaults and configure the measurement board to the desired measurement. Follow the Reset Analyzer menu (Fig. 5­5) to reconfigure the analyzer to display new measurements or measurement units. 2. Program menus to adjust any of the programmable configuration items. Use the following configuration and programming guidelines for the applicable measurement. 41 MODEL 1056 SECTION 6.0 PROGRAMMING THE MEASUREMENTS 6.2 pH MEASUREMENT PROGRAMMING 6.2.1 Description This section describes how to configure the Model 1056 analyzer for pH measurements. The following programming and configuration functions are covered. TABLE 6­1. pH Measurement Programming Measure pH default setting Description Sec. Menu function: 6.2.2 Measurement type: 6.2.3 Preamp location: 6.2.4 Solution temperature correction Off Select Off, ultra­pure, high pH, custom 6.2.5 Temp coefficient Enter the temp coefficient 6.2.6 Resolution: 6.2.7 Filter: 6.2.8 Reference Z: pH Analyzer (custom) 0.01pH 4 sec Low Select pH, ORP, Redox, Ammonia, Fluoride, Custom ISE Identify preamp location Select 0.01pH or 0.1pH for pH display resolution Override the default input filter, enter 0­999 seconds Select low or high reference impedance A detailed flow diagram for pH programming is provided at the end of Sec. 6 to guide you through all basic programming and configuration functions. To configure the pH measurement board: 1. Press MENU 2. Scroll down to Program. Press ENTER. 3. Scroll down to Measurement. Press ENTER. 4. Select Sensor 1 or Sensor 2 corresponding to pH. Press ENTER. The adjacent screen format will appear (factory defaults are shown). To program any function, scroll to the desired item and press ENTER. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Configure Measure: pH Preamp: Analyzer Sol’n Temp Corr: Off T Coeff: ­0.029pH/°C Resolution: 0.01pH Filter: 4 sec Reference Z: Low The following sub­sections provide you with the initial display screen that appears for each configuration function. Use the flow diagram for pH programming at the end of Sec. 6 and the Model 1056 live screen prompts for each function to complete configuration and programming. 6.2.2 Measurement The display screen for selecting the measurement is shown. The default value is displayed in bold type. Refer to the pH/ORP Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Measurement pH ORP Redox Ammonia Fluoride Custom ISE 6.2.3 Preamp The display screen for identifying the Preamp location is shown. The default value is displayed in bold type. Refer to the pH/ORP Programming flow diagram to complete this function. 42 S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Preamp Analyzer Sensor/JBox MODEL 1056 SECTION 6.0 PROGRAMMING THE MEASUREMENTS 6.2.4 Solution Temperature Correction The display screen for selecting the Solution temperature correction algorithm is shown. The default value is displayed in bold type. Refer to the pH/ORP Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Sol’n Temp Corr. Off Ultra Pure Water High pH Custom 6.2.5 Temperature Coefficient The display screen for entering the custom solution tem­ perature coefficient is shown. The default value is dis­ played in bold type. Refer to the pH/ORP Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Sol’n Temp Coeff. ­ 0.032pH/ºC 6.2.6 Resolution The display screen for selecting 0.01pH or 0.1pH for pH display resolution is shown. The default value is displayed in bold type. Refer to the pH/ORP Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Resolution 0.01pH 0.1pH 6.2.7 Filter The display screen for entering the input filter value in seconds is shown. The default value is displayed in bold type. Refer to the pH/ORP Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Input filter 04 sec 6.2.8 Reference Impedence The display screen for selecting Low or High Reference impedance is shown. The default value is displayed in bold type. Refer to the pH/ORP Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Reference Z Low High 6.3 ORP MEASUREMENT PROGRAMMING 6.3.1 Description The section describes how to configure the Model 1056 analyzer for ORP measurements. The following programming and configuration functions are covered: TABLE 6­2. ORP Measurement Programming Measure ORP Sec. Menu function: 6.3.2 Measurement type: 6.3.3 Preamp location: 6.3.4 Filter: 6.3.5 Reference Z: default pH Analyzer 4 sec Low Description Select pH, ORP, Redox, Ammonia, Fluoride, Custom ISE Identify preamp location Override the default input filter, enter 0­999 seconds Select low or high reference impedance 43 MODEL 1056 A detailed flow diagram for ORP programming is provided at the end of Sec. 6 to guide you through all basic programming and configuration functions. To configure the ORP measurement board: 1. Press MENU 2. Scroll down to Program. Press ENTER. 3. Scroll down to Measurement. Press ENTER. 4. Select Sensor 1 or Sensor 2 corresponding to ORP. Press ENTER. The adjacent screen format will appear (factory defaults are shown). To program any displayed function, scroll to the desired item and press ENTER. SECTION 6.0 PROGRAMMING THE MEASUREMENTS S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Configure Measure: pH Preamp: Analyzer Flter: 4 sec Reference Z: Low The following sub­sections provide you with the initial display screen that appears for each configuration function. Use the flow diagram for ORP programming at the end of Sec. 6 and the Model 1056 live screen prompts for each function to complete configuration and programming. 6.3.2 Measurement The display screen for selecting the measurement is shown. The default value is displayed in bold type. Refer to the pH/ORP Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC 6.3.3 Preamp The display screen for identifying the Preamp location is shown. The default value is displayed in bold type. Refer to the pH/ORP Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC 6.3.4 Filter The display screen for entering the input filter value in seconds is shown. The default value is displayed in bold type. Refer to the pH/ORP Programming flow diagram to complete this function. 6.3.5 Reference Impedence The display screen for Selecting Low or high Reference impedance is shown. The default value is displayed in bold type. Refer to the pH/ORP Programming flow diagram to complete this function. 44 SN Measurement pH ORP Redox Ammonia Fluoride Custom ISE SN Preamp Analyzer Sensor/JBox S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Input filter 04 sec S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Reference Z Low High MODEL 1056 SECTION 6.0 PROGRAMMING THE MEASUREMENTS 6.4 CONTACTING CONDUCTIVITY MEASUREMENT PROGRAMMING 6.4.1 Description The section describes how to configure the Model 1056 analyzer for conductivity measurements using contacting conductivity sensors. The following programming and configuration functions are covered. TABLE 6­3. Contacting Conductivity Measurement Programming Measure Contacting Conductivity default Description Sec. Menu function: 6.4.2 Type: 6.4.3 Measure: 6.4.4 Range: Auto 6.4.5 Cell K: 1.00000/cm 6.4.6 RTD Offset: 0.00ºC Enter the RTD Offset 6.4.7 RTD Slope: 0 Enter the RTD Slope 6.4.8 Temp Comp: 6.4.9 Slope: 6.4.10 Ref Temp: 6.4.11 Filter: 6.4.12 Custom 6.4.13 Cal Factor: 2­Electrode Conductivity Slope 2.00%/°C 25.0°C 2 sec Setup 0.95000/cm Select 2­Electrode or 4­Electrode type sensors Select Conductivity, Resistivity, TDS. Salinity or % conc Select measurement Auto­range or specific range Enter the cell Constant for the sensor Select Temp Comp: Slope, Neutral Salt, Cation or Raw Enter the linear temperature coefficient Enter the Reference temp Override the default input filter, enter 0­999 seconds Enter 2­5 data points in ppm and µS/cm for custom curves Enter the Cal Factor for 4­Electrode sensors from the sensor tag A detailed flow diagram for contacting conductivity programming is provided at the end of Sec. 6 to guide you through all basic programming and configuration functions. To configure the contacting conductivity measurement board: 1. Press MENU 2. Scroll down to Program. Press ENTER. 3. Scroll down to Measurement. Press ENTER. 4. Select Sensor 1 or Sensor 2 corresponding to contacting conductivity. Press ENTER. The adjacent screen format will appear (factory defaults are shown). To program any displayed function, scroll to the desired item and press ENTER. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Configure Type: 2­Electrode Measure: Cond Range: Auto Cell K: 1.00000/cm RTD Offset: 0.00ºC RTD Slope: 0 Temp Comp: Slope Slope: 2.00%/°C Ref Temp: 25.0°C Filter: 2 sec Custom Setup The following sub­sections provide you with the initial display screen that appears for each configuration function. Use the flow diagram for contacting conductivity programming at the end of Sec. 6 and the Model 1056 live screen prompts for each function to complete configuration and programming. 6.4.2 Sensor Type The display screen for selecting 2­Electrode or 4­Electrode type sensors is shown. The default value is displayed in bold type. Refer to the contacting conductivity Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Type 2­Electrode 4­Electrode 45 MODEL 1056 6.4.3 Measure The display screen for selecting the measurement is shown. The default value is displayed in bold type. Refer to the contacting conductivity Programming flow diagram to complete this function. SECTION 6.0 PROGRAMMING THE MEASUREMENTS S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Measurement Conductivity Resistivity TDS Salinity NaOH (0­12%) HCl (0­15%) Low H2SO4 High H2SO4 NaCl (0­20%) Custom Curve 6.4.4 Range The display screen for Selecting Auto­ranging or a specific range is shown. The default value is displayed in bold type. Note: Ranges are shown as conductance, not conductivity. Refer to the contacting conductivity Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC 6.4.5 Cell Constant The display screen for entering a cell Constant for the sensor is shown. The default value is displayed in bold type. Refer to the contacting conductivity Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC 6.4.6 RTD Offset The display screen for Entering the RTD Offset for the sensor is shown. The default value is displayed in bold type. Refer to the contacting conductivity Programming flow diagram to complete this function. SN Range Auto 50 µS 500 µS 2000 µS 20 mS 200 mS 600 mS SN Cell Constant 1.00000 /cm S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN RTD Offset 0.00°C 6.4.7 RTD Slope The display screen for entering the RTD slope for the sensor is shown. The default value is displayed in bold type. Refer to the contacting conductivity Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC 6.4.8 Temp Comp The display screen for Selecting Temperature Compensation as Slope, Neutral Salt, Cation or Raw is shown. The default value is displayed in bold type. Refer to the contacting conductivity Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC 46 SN RTD Slope 2.00%/ºC SN Temp Comp Slope Neutral Salt Cation Raw MODEL 1056 6.4.9 Slope The display screen for Entering the conductivity/temp Slope is shown. The default value is displayed in bold type. Refer to the contacting conductivity Programming flow diagram to complete this function. 6.4.10 Reference Temp The display screen for manually entering the Reference temperature is shown. The default value is displayed in bold type. Refer to the contacting conductivity Programming flow diagram to complete this function. 6.4.11 Filter The display screen for entering the input filter value in seconds is shown. The default value is displayed in bold type. Refer to the contacting conductivity Programming flow diagram to complete this function. 6.4.12 Custom Setup The display screens for creating a custom curve for converting conductivity to concentration is shown. Refer to the contacting conductivity Programming flow diagram to complete this function. When the custom curve data entry is complete, press ENTER. The display will confirm the determination of a custom curve fit to the entered data by displaying this screen: If the custom curve fit is not completed or is unsuccessful, the display will read as follows and the screen will return to the beginning custom curve screen. 6.4.13 Cal Factor Upon initial installation and power up, if 4­electrode was selected for the sensor type in the Quick Start menus, the user enters a Cell Constant and a “Cal Factor” using the instrument keypad. The cell constant is needed to convert measured conductance to conductivity as displayed on the analyzer screen. The “Cal Factor” entry is needed increase the accuracy of the live conductivity readings, especially at low conduc­ tivity readings below 20uS/cm. Both the Cell Constant and the “Cal Factor” are printed on the tag attached to the 4­electrode sensor/cable. SECTION 6.0 PROGRAMMING THE MEASUREMENTS S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Slope 2.00 %/ºC S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Ref Temp (25.0ºC normal) +25.0ºC S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Input filter 02 sec S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Custom Curve Configure Enter Data Points Calculate Curve S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calculate Curve Custom curve fit completed. In Process Cal recommended. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calculate Curve Failure The display screen for entering Cal Factor is shown. The default value is displayed in bold type. If necessary after initial installation and start­up, enter the “Cal Factor” as printed on the sensor tag. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Cal Factor 0.95000/cm 47 MODEL 1056 SECTION 6.0 PROGRAMMING THE MEASUREMENTS 6.5 TOROIDAL CONDUCTIVITY MEASUREMENT PROGRAMMING 6.5.1 Description The section describes how to configure the Model 1056 analyzer for conductivity measurements using inductive/toroidal sensors. The following programming and configuration functions are covered. TABLE 6­4. Toroidal Conductivity Measurement Programming Measure Toroidal conductivity Sec. Menu function: default 6.5.2 Model: 6.5.3 Measure: 6.5.4 Range: Auto 6.5.5 Cell K: 3.00000/cm 6.5.6 Temp Comp: 6.5.7 Slope: 6.5.8 Ref Temp: 6.5.9 Filter: 6.5.10 Custom 228 Conductivity Slope 2.00%/°C 25.0°C 2 sec Setup Description Select sensor type Select Conductivity, Resistivity, TDS, Salinity or % conc Select measurement Auto­range or specific range Enter the cell Constant for the sensor Select Temp Comp: Slope, Neutral Salt, or Raw Enter the linear temperature coefficient Enter the Reference temp Override the default input filter, enter 0­999 seconds Enter 2­5 data points in ppm and µS/cm for custom curves A detailed flow diagram for toroidal conductivity programming is provided at the end of Sec. 6 to guide you through all basic programming and configuration functions. To configure the toroidal conductivity measurement board: 1. Press MENU 2. Scroll down to Program. Press ENTER. 3. Scroll down to Measurement. Press ENTER. 4. Select Sensor 1 or Sensor 2 corresponding to toroidal conductivity. Press ENTER. The adjacent screen format will appear (factory defaults are shown). To program any displayed function, scroll to the desired item and press ENTER. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Configure Model: 228 Measure: Cond Range: Auto Cell K: 3.00000/cm RTD Offset: 0.00ºC RTD Slope: 0 Temp Comp: Slope Slope: 2.00%/°C Ref Temp: 25.0°C Filter: 2 sec Custom Setup The following sub­sections provide you with the initial display screen that appears for each configuration function. Use the flow diagram for toroidal conductivity programming at the end of Sec. 6 and the Model 1056 live screen prompts for each function to complete configuration and programming. 6.5.2 Sensor Model The display screen for selecting the sensor model is shown. The default value is displayed in bold type. Refer to the toroidal conductivity Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Model 228 225 226 247 Other 48 MODEL 1056 6.5.3 Measure The display screen for selecting the measurement is shown. The default value is displayed in bold type. Refer to the toroidal conductivity Programming flow diagram to complete this function. SECTION 6.0 PROGRAMMING THE MEASUREMENTS S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Measurement Conductivity Resistivity TDS Salinity NaOH (0­12%) HCl (0­15%) Low H2SO4 High H2SO4 NaCl (0­20%) Custom Curve 6.5.4 Range The display screen for Selecting Auto­ranging or a specific range is shown. The default value is displayed in bold type. Note: Ranges are shown as conductance, not conductivity. Refer to the toroidal conductivity Programming flow diagram to complete this function. 6.5.5 Cell Constant The display screen for entering a cell Constant for the sensor is shown. The default value is displayed in bold type. Refer to the toroidal conductivity Programming flow diagram to complete this function. 6.5.6 Temp Comp The display screen for Selecting Temperature Compensation as Slope, Neutral Salt, or Raw is shown. The default value is displayed in bold type. Refer to the toroidal conductivity Programming flow diagram to com­ plete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Range Auto 2000 mS 50 mS 2 mS 200µS S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Cell Constant 3.00000 /cm S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Temp Comp Slope Neutral Salt Raw 49 MODEL 1056 6.5.7 Slope The display screen for Entering the conductivity/temp Slope is shown. The default value is displayed in bold type. Refer to the toroidal conductivity Programming flow diagram to complete this function. SECTION 6.0 PROGRAMMING THE MEASUREMENTS S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Slope 2.00%/ºC 6.5.8 Ref Temp The display screen for manually Entering the Reference temperature is shown. The default value is displayed in bold type. Refer to the toroidal conductivity Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC 6.5.9 Filter The display screen for entering the input filter value in seconds is shown. The default value is displayed in bold type. Refer to the toroidal conductivity Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC 6.5.10 Custom Setup The display screens for creating custom curves for con­ verting conductivity to concentration is shown. Refer to the toroidal conductivity Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC When the custom curve data entry is complete, press ENTER. The display will confirm the determination of a custom curve fit to the entered data by displaying this screen: If the custom curve fit is not completed or is unsuccessful, the display will read as follows and the screen will return to the beginning custom curve screen. 50 SN Ref Temp (25.0ºC normal) +25.0ºC SN Input filter 02 sec SN Custom Curve Configure Enter Data Points Calculate Curve S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calculate Curve Custom curve fit completed. In Process Cal recommended. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calculate Curve Failure MODEL 1056 SECTION 6.0 PROGRAMMING THE MEASUREMENTS 6.6 CHLORINE MEASUREMENT PROGRAMMING With a Chlorine measurement board installed, Model 1056 can measure any of four variants of Chlorine: • Free Chlorine • Total Chlorine • Monochloramine • pH­independent Free Chlorine The section describes how to configure the Model 1056 analyzer for Chlorine measurements. 6.6.1 FREE CHLORINE MEASUREMENT PROGRAMMING 6.6.1.1 Description This Chlorine sub­section describes how to configure the Model 1056 analyzer for Free Chlorine measurement using amperometric chlorine sensors. The following programming and configuration functions are covered: TABLE 6­5. Free Chlorine Measurement Programming Measure Free Chlorine Sec. Menu function: default Free Chlorine Description Select Free Chlorine, pH Ind. Free Cl. Total Cl, Monochloramine 6.6.1.2 Measure: 6.6.1.3 Units: ppm Select units ppm or mg/L 6.6.1.4 Filter: 5sec Override the default input filter, enter 0­999 seconds 6.6.1.5 Free Cl Correct: 6.6.1.6 Manual pH: 7.00 pH 6.6.1.7 Resolution: 0.001 Live Select Live/Continuous pH correction or Manual For Manual pH correction, enter the pH value Select display resolution 0.01 or 0.001 A detailed flow diagram for programming of all chlorine measurements is provided at the end of Sec. 6 to guide you through all basic programming and configuration functions. To configure the chlorine measurement board for free chlorine: 1. Press MENU 2. Scroll down to Program. Press ENTER. 3. Scroll down to Measurement. Press ENTER. 4. Select Sensor 1 or Sensor 2 corresponding to chlorine. Press ENTER. The adjacent screen format will appear (factory defaults are shown). To program any displayed function, scroll to the desired item and press ENTER. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Configure Measure: Free Chlorine Units: ppm Filter: 5sec Free Cl Correct: Live Manual pH: 7.00 pH Resolution: 0.001 51 MODEL 1056 SECTION 6.0 PROGRAMMING THE MEASUREMENTS The following sub­sections provide you with the initial display screen that appears for each configuration function. Use the flow diagram for chlorine programming at the end of Sec. 6 and the Model 1056 live screen prompts for each function to complete configuration and programming. 6.6.1.2 Measure The display screen for selecting the measurement is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. 6.6.1.3 Units The display screen for selecting units as ppm or mg/L is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. 6.6.1.4 Filter The display screen for entering the input filter value in seconds is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. 6.6.1.5 Free Chlorine pH Correction The display screen for Selecting Live/Continuous pH correction or Manual pH correction is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. 6.6.1.6 Manual pH Correction The display screen for manually entering the pH value of the measured process liquid is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. 6.6.1.7 Resolution The display screen for selecting display resolution as 0.001 or 0.01 is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. 52 S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Measurement Free Chlorine pH Independ. Free Cl Total Chlorine Monochloramine S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Units ppm mg/L S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Input filter 05 sec S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Free Cl pH Correction Live/Continuous Manual S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Manual pH 07.00 pH S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Resolution ­ 0.001 0.01 MODEL 1056 SECTION 6.0 PROGRAMMING THE MEASUREMENTS 6.6.2 TOTAL CHLORINE MEASUREMENT PROGRAMMING 6.6.2.1 Description This Chlorine sub­section describes how to configure the Model 1056 analyzer for Total Chlorine measurement using amperometric chlorine sensors. The following programming and configuration functions are covered: TABLE 6­6. Total Chlorine Measurement Programming Measure Total Chlorine Sec. Menu function: default 6.6.2.2 Measure: 6.6.2.3 Units: ppm 6.6.2.4 Filter: 5sec 6.6.2.5 Resolution: 0.001 Free Chlorine Description Select Free Chlorine, pH Ind. Free Cl. Total Cl, Monochloramine Select units ppm or mg/L Override the default input filter, enter 0­999 seconds Select 0.01 or 0.001 display resolution A detailed flow diagram for programming of all chlorine measurements is provided at the end of Sec. 6 to guide you through all basic programming and configuration functions. To configure the chlorine measurement board for total chlorine: 1. Press MENU 2. Scroll down to Program. Press ENTER. 3. Scroll down to Measurement. Press ENTER. 4. Select Sensor 1 or Sensor 2 corresponding to chlorine. Press ENTER. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Configure Measure: Free Chlorine Units: ppm Filter: 5sec Resolution: 0.001 The adjacent screen format will appear (factory defaults are shown). To program any displayed function, scroll to the desired item and press ENTER. The following sub­sections provide you with the initial display screen that appears for each configuration function. Use the flow diagram for chlorine programming at the end of Sec. 6 and the Model 1056 live screen prompts for each function to complete configuration and programming. 6.6.2.2 Measure The display screen for selecting the measurement is shown. The default value is displayed in bold type. Refer to the chlorine Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC 6.6.2.3 Units The display screen for selecting units as ppm or mg/L is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Measurement Free Chlorine pH Independ. Free Cl Total Chlorine Monochloramine SN Units ppm mg/L 53 MODEL 1056 SECTION 6.0 PROGRAMMING THE MEASUREMENTS 6.6.2.4 Filter The display screen for entering the input filter value in seconds is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Input filter 05 sec 6.6.2.5 Resolution The display screen for selecting display resolution as 0.001 or 0.01 is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Resolution 0.001 0.01 6.6.3 MONOCHLORAMINE MEASUREMENT PROGRAMMING 6.6.3.1 Description This Chlorine sub­section describes how to configure the Model 1056 analyzer for Monochloramine measurement using amperometric chlorine sensors. The following programming and configuration functions are covered: TABLE 6­7. Monochloramine Measurement Programming Measure Sec. Monochloramine 6.6.3.2 6.6.3.3 6.6.3.4 6.6.3.5 Menu function: default Measure: Free Chlorine Units: ppm Filter: 5sec Resolution: 0.001 Description Select Free Chlorine, pH Ind. Free Cl. Total Cl, Monochloramine Select units ppm or mg/L Override the default input filter, enter 0­999 seconds Select 0.01pH or 0.1ppm/mg/L for display Resolution A detailed flow diagram for programming of all chlorine measurements is provided at the end of Sec. 6 to guide you through all basic programming and configuration functions. To configure the chlorine measurement board for monochloramine: 1. Press MENU 2. Scroll down to Program. Press ENTER. 3. Scroll down to Measurement. Press ENTER. 4. Select Sensor 1 or Sensor 2 corresponding to chlorine. Press ENTER. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Configure Measure: Free Chlorine Units: ppm 5sec Filter: Resolution: 0.001 The following screen format will appear (factory defaults are shown). To program any displayed function, scroll to the desired item and press ENTER. The following sub­sections provide you with the initial display screen that appears for each configuration function. Use the flow diagram for chlorine programming at the end of Sec. 6 and the Model 1056 live screen prompts for each function to complete configuration and programming. 6.6.3.2­Measure: Monochloramine The display screen for selecting the measurement is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. 54 S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Measurement Free Chlorine pH Independ. Free Cl Total Chlorine Monochloramine MODEL 1056 SECTION 6.0 PROGRAMMING THE MEASUREMENTS 6.6.3.3 Units The display screen for selecting units as ppm or mg/L is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. 6.6.3.4 Filter The display screen for entering the input filter value in seconds is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. 6.6.3.5 Resolution The display screen for selecting display resolution as 0.001 or 0.01 is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Units ppm mg/L S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Input filter 05 sec S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Resolution 0.001 0.01 6.6.4 pH­INDEPENDENT FREE CHLORINE MEASUREMENT PROGRAMMING 6.6.4.1 Description This Chlorine sub­section describes how to configure the Model 1056 analyzer for Free Chlorine measurements using the pH­independent free chlorine sensor, Model 498CL­01, manufactured by Rosemount Analytical. The following programming and configuration functions are covered: TABLE 6­8. pH­independent Free Chlorine Measurement Programming Measure pH­independent Free Chlorine Sec. 6.6.4.2 Menu function: default Measure: pH Indep Free Cl 6.6.4.3 Units: ppm 6.6.4.4 Filter: 5sec 6.6.4.5 Resolution: 0.001 Description Select Free Chlorine, pH Ind. Free Cl. Total Cl, Monochloramine Select units ppm or mg/L Override the default input filter, enter 0­999 seconds Select 0.01pH or 0.1ppm/mg/L for display Resolution A detailed flow diagram for programming of all chlorine measurements is provided at the end of Sec. 6 to guide you through all basic programming and configuration functions. To configure the chlorine measurement board for pH­independent free chlorine: 1. Press MENU 2. Scroll down to Program. Press ENTER. 3. Scroll down to Measurement. Press ENTER. 4. Select Sensor 1 or Sensor 2 corresponding to chlorine. Press ENTER. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Configure Measure: Free Chlorine Units: ppm Filter: 5sec Resolution: 0.001 The adjacent screen format will appear (factory defaults are shown). To program any displayed function, scroll to the desired item and press ENTER. 55 MODEL 1056 SECTION 6.0 PROGRAMMING THE MEASUREMENTS The following sub­sections provide you with the initial display screen that appears for each configuration function. Use the flow diagram for chlorine programming at the end of Sec. 6 and the Model 1056 live screen prompts for each function to complete configuration and programming. 6.6.4.2 Measurement: pH­independent Free Chlorine The display screen for selecting the measurement is shown. The default value is displayed in bold type. Refer to the chlorine Programming flow diagram to complete this function. 6.6.4.3 Units The display screen for selecting units as ppm or mg/L is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. 6.6.4.4 Filter The display screen for entering the input filter value in seconds is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. 6.6.4.5 Resolution The display screen for selecting display resolution as 0.001 or 0.01 is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. 56 S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Measurement Free Chlorine pH Independ. Free Cl Total Chlorine Monochloramine S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Units ppm mg/L S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Input filter 05 sec S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Resolution 0.001 0.01 MODEL 1056 SECTION 6.0 PROGRAMMING THE MEASUREMENTS 6.7 OXYGEN MEASUREMENT PROGRAMMING 6.7.1 Description This section describes how to configure the Model 1056 analyzer for dissolved and gaseous oxygen measurement using amperometric oxygen sensors. The following programming and configuration functions are covered: TABLE 6­9. Oxygen Measurement Programming Measure Oxygen Sec. Menu function: default Description 6.7.2 Type: Water/Waste Select Water/Waste, Trace. BioRx, BioRx­Other, Brew, %O2 In Gas 6.7.3 Units: ppm Select ppm, mg/L, ppb, µg/L, % Sat, %O2­Gas, ppm Oxygen­Gas 6.7.4 Partial Press: mmHg Select mm Hg, in Hg. atm, kPa, mbar or bar for Partial pressure 6.7.5 Salinity: 00.0‰ Enter Salinity as ‰ 6.7.6 Filter: 6.7.7 Pressure Units: 6.7.8 Use Press: 5sec Override the default input filter, enter 0­999 seconds bar Select pressure units: mm Hg, in Hg,. Atm, kPa, mbar, bar At Air Cal Select atmospheric pressure source – internal or mA Input A detailed flow diagram for oxygen programming is provided at the end of Sec. 6 to guide you through all basic pro­ gramming and configuration functions. To configure the Oxygen measurement board: 1. Press MENU 2. Scroll down to Program. Press ENTER. 3. Scroll down to Measurement. Press ENTER. 4. Select Sensor 1 or Sensor 2 corresponding to Oxygen. Press ENTER. The adjacent screen format will appear (factory defaults are shown). To program any displayed function, scroll to the desired item and press ENTER. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Configure Type: Water/Waste Units: ppm Partial Press: mmHg Salinity: 00.0‰ Filter: 5sec Pressure Units: bar Use Press: At Air Cal Custom Setup The following sub­sections show the initial display screen that appears for each configuration function. Use the flow diagram for oxygen programming at the end of Sec. 6 and the Model 1056 live screen prompts for each function to complete configuration and programming. 6.7.2 Oxygen Measurement application The display screen for programming the measurement is shown. The default value is displayed in bold type. Refer to the Oxygen Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Type Water/Waste Trace Oxygen BioRx­Rosemount BioRx­Other Brewing Oxygen In Gas 6.7.3 Units The display screen for selecting units as ppm , mg/L, ppb, µg/L, % Saturation, %Oxygen in Gas, or ppm Oxygen in Gas is shown. The default value is displayed in bold type. Refer to the Oxygen Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Units ppm mg/L ppb µg/L % Saturation Partial Pressure % Oxygen In Gas ppm Oxygen In Gas 57 MODEL 1056 6.7.4 Partial Press The display screen for selecting pressure units for Partial pressure is shown. This selection is needed if the specified measurement is Partial pressure. The default value is displayed in bold type. Refer to the Oxygen Programming flow diagram to complete this function. 6.7.5 Salinity The display screen for Entering the Salinity (as parts per thousand) of the process liquid to be measured is shown. The default value is displayed in bold type. Refer to the Oxygen Programming flow diagram to complete this function. Enter Salinity as ‰ 6.7.6 Filter The display screen for entering the input filter value in seconds is shown. The default value is displayed in bold type. Refer to the Oxygen Programming flow diagram to complete this function. 6.7.7 Pressure Units The display screen for selecting pressure units for atmospheric pressure is shown. This selection is needed for the display of atmospheric pressure measured by the onboard pressure transducer on the Oxygen measurement board. The default value is displayed in bold type. Refer to the Oxygen Programming flow diagram to complete this function. 6.7.8 Use Pressure The display screen for selecting atmospheric pressure source. The default value is displayed in bold type. Refer to the Oxygen Programming flow diagram to com­ plete this function. 58 SECTION 6.0 PROGRAMMING THE MEASUREMENTS S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Partial Press mm Hg in Hg atm kPa mbar bar S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Salinity 00.0 ‰ S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Input filter 05 sec S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC Pressure Units mm Hg in Hg atm kPa mbar bar S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Use Pressure? At Air Cal mA Input MODEL 1056 SECTION 6.0 PROGRAMMING THE MEASUREMENTS 6.8 OZONE MEASUREMENT PROGRAMMING 6.8.1 Description This section describes how to configure the 1056 analyzer for ozone measurement using amperometric ozone sen­ sors. The following programming and configuration functions are covered: TABLE 6­10. Ozone Measurement Programming Measure Ozone Sec. Menu function: default 6.8.2 Units: ppm 6.8.3 Filter: 5sec 6.8.4 Resolution: 0.001 Description Select units ppm, mg/L, ppb, µg/L Override the default input filter, enter 0­999 seconds Select 0.01or 0.001 for display resolution A detailed flow diagram for ozone programming is provided at the end of Sec. 6 to guide you through all basic programming and configuration functions. To configure the Ozone measurement board: 1. Press MENU 2. Scroll down to Program. Press ENTER. 3. Scroll down to Measurement. Press ENTER. 4. Select Sensor 1 or Sensor 2 corresponding to Ozone. Press ENTER. The adjacent screen format will appear (factory defaults are shown). To program any displayed function, scroll to the desired item and press ENTER. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Configure Units: ppm Filter: 5 sec Resolution: 0.001 The following sub­sections show the initial display screen that appears for each configuration function. Use the flow diagram for ozone programming at the end of Sec. 6 and the Model 1056 live screen prompts for each function to complete configuration and programming. Note: Ozone measurement boards are detected automatically by the analyzer. No measurement selection is necessary. 6.8.2 Units The display screen for selecting measurement units is shown. The default value is displayed in bold type. Refer to the Ozone Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC 6.8.3 Filter The display screen for entering the input filter value in seconds is shown. The default value is displayed in bold type. Refer to the Ozone Programming flow diagram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC 6.8.4 Resolution The display screen for selecting display resolution as 0.001 or 0.01 is shown. The default value is displayed in bold type. Refer to the Ozone Programming flow diagram to complete this function. SN Units ppm mg/L ppb µg/L SN Input filter 05 sec S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Resolution 0.001 0.01 59 MODEL 1056 SECTION 6.0 PROGRAMMING THE MEASUREMENTS 6.9 TURBIDITY MEASUREMENT PROGRAMMING 6.9.1 DESCRIPTION This section describes how to configure the Model 1056 analyzer for Turbidity measurements. The following programming and configuration functions are covered. TABLE 6­11 TURBIDITY MEASUREMENT PROGRAMMING Measure Turbidity Sec. Menu function: 6.9.2 Measurement type: Turbidity 6.9.3 Measurement units: NTU 6.9.4 Enter TSS* Data: 6.9.5 Filter: 6.9.6 Bubble Rejection: default Description Select Turbidity or TSS calculation (estimated TSS) NTU, FTU, FNU Enter TSS and NTU data to calculate TSS based on Turbidity 20 sec On Override the default input filter, enter 0­999 seconds Intelligent software algorithm to eliminate erroneous readings caused by bubble accumulation in the sample *TSS: Total Suspended Solids A detailed flow diagram for Turbidity programming is provided at the end of Sec. 6 to guide you through all basic programming and configuration functions. To 1. 2. 3. 4. configure the Turbidity measurement board: Press MENU Scroll down to Program. Press ENTER. Scroll down to Measurement. Press ENTER. Select Sensor 1 or Sensor 2 corresponding to Turbidity. Press ENTER. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Configure Measure: Turbidity Units: NTU Enter TSS Data Filter: 20sec Bubble Rejection: On The following screen format will appear (factory defaults are shown). To program Turbidity, scroll to the desired item and press ENTER. The following sub­sections provide you with the initial display screen that appears for each programming routine. Use the flow diagram for Turbidity programming at the end of Sec. 6 and the live screen prompts to complete programming. 6.9.2 Measurement The display screen for selecting the measurement is shown. The default measurement is displayed in bold type. Refer to the Turbidity Programming flow diagram to complete this function. 6.9.3 Units The display screen for selecting the measurement units is shown. The default value is displayed in bold type. Refer to the Turbidity Programming flow diagram to complete this function. 60 S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Measurement Turbidity Calculated TSS S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Units NTU FTU FNU MODEL 1056 SECTION 6.0 PROGRAMMING THE MEASUREMENTS If TSS data (Total Suspended Solids) calculation is selected, the following screen will be displayed. Refer to the Turbidity programming flow diagram to complete this function. 6.9.4 Enter TSS Data The display screen for entering TSS Data is shown. The default values are displayed. Refer to the Turbidity Programming flow diagram to complete this function S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Units ppm mg/L none S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN TSS Data Pt1 TSS: 0.000ppm Pt1 Turbid: 0.000NTU Pt2 TSS: 100.0ppm Pt2 Turbid: 100.0NTU Calculate Note: Based on user­entered NTU data, calculating TSS as a straight line curve could cause TSS to go below zero. The following screen lets users know that TSS will become zero below a certain NTU value. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN TSS Data Calculation Complete Calculated TSS = 0 below xxxx NTU The following illustration shows the potential for calculated TSS to go below zero Normal case: TSS is always a positive number when Turbidity is a positive number. Abnormal case: TSS can be a negative number when Turbidity is a positive number. 61 MODEL 1056 When the TSS data entry is complete, press ENTER. The display will confirm the determination of a TSS straight line curve fit to the entered NTU/turbidity data by displaying this screen: The following screen may appear if TSS calculation is unsuccessful. Re­entry of NTU and TSS data is required. SECTION 6.0 PROGRAMMING THE MEASUREMENTS S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN TSS Data Calculation Complete S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN TSS Data Data Entry Error Press EXIT 6.9.5 Filter The display screen for entering the input filter value in seconds is shown. The default value is displayed in bold type. Refer to the Turbidity Programming flow diagram to complete this function. 6.9.6 Bubble Rejection Bubble rejection is an internal software algorithm that characterizes turbidity readings as bubbles as opposed to true turbidity of the sample. With Bubble rejection enabled, these erroneous readings are elimi­ nated from the live measurements shown on the dis­ play and transmitted via the current outputs. The display screen for selecting bubble rejection algo­ rithm is shown. The default setting is displayed in bold type. Refer to the Turbidity Programming flow diagram to complete this function. 62 S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Input Filter 020sec S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Bubble Rejection On Off MODEL 1056 SECTION 6.0 PROGRAMMING THE MEASUREMENTS 6.10 FLOW MEASUREMENT PROGRAMMING 6.10.1 DESCRIPTION This section describes how to configure the 1056 analyzer for flow measurement when used with a compatible pulse flow sensor. The following programming and configuration functions are covered. TABLE 6­12 FLOW MEASUREMENT PROGRAMMING Measure Flow To 1. 2. 3. 4. Sec. Menu function: 6.10.2 Measurement type 6.10.3 Measurement units: 6.10.4 Enter TSS* Data: default Pulse Flow GPH 0 Sec Description Select Pulse Flow or mA Current Input Select GPM, GPH, cu ft/min, cu ft/hour, LPM, L/hour, m3/hr Override the default input filter, enter 0­999 seconds configure the flow measurement board: Press MENU Scroll down to Program. Press ENTER. Scroll down to Measurement. Press ENTER. Select Sensor 1 or Sensor 2 corresponding to flow. Press ENTER. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Configure Measure: Pulse Flow Units: GPM Filter: 5sec The following screen format will appear (factory defaults are shown). To program pulse flow, scroll to the desired item and press ENTER. The following sub­sections provide you with the initial display screen that appears for each programming routine. Use the diagram for pulse flow programming at the end of Sec. 6 and the live screen prompts to complete programming. 6.10.2 Measurement The display screen for selecting the measurement is shown. The default measurement is displayed in bold type. Refer to the pulse flow Programming diagram to complete this function. 6.10.3 Units The display screen for selecting measurement units is shown. The default units are displayed in bold type. Refer to the pulse flow Programming diagram to com­ plete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Measurement Pulse Flow mA Input S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Units GPM GPH cu ft/min cu ft/hour L/min L/hour m3/hour 6.10.4 Filter The display screen for entering the input filter value in seconds is shown. The default value is displayed in bold type. Refer to the pulse flow Programming dia­ gram to complete this function. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Input Filter 005sec 63 MODEL 1056 SECTION 6.0 PROGRAMMING THE MEASUREMENTS 6.11 CURRENT INPUT PROGRAMMING 6.11.1 DESCRIPTION This section describes how to configure the Model 1056 analyzer for current input measurement when wired to an external device that transmits 4­20mA or 0­20mA analog current output. The following programming and configu­ ration functions are covered. TABLE 6­13 CURRENT INPUT PROGRAMMING Measure Current Input Sec. Menu function: 6.11.2 Measurement type 6.11.3 mA Input default mA input Temperature ºC Description Override the default (Flow) and select mA current input Select Temperature, Pressure, Flow or Other 6.11.4 Measurement units: 6.11.5 Input Range: 4­20mA Select 4­20mA or 0­20mA 6.11.6 Low Value: 0.000ºC Enter the low measurement value to assign to 4mA 6.11.7 High Value: 100.0ºC Enter the high measurement value to assign to 20mA 6.11.8 Filter: 05 sec Override the default input filter, enter 0­999 seconds Select measurement units based on selected input device type A detailed flow diagram for current input programming is provided at the end of Sec. 6 to guide you through all basic programming and configuration functions. To 1. 2. 3. 4. configure the current input measurement board: Press MENU Scroll down to Program. Press ENTER. Scroll down to Measurement. Press ENTER. Select Sensor 1 or Sensor 2 corresponding to current input. Press ENTER. Note that factory default is Pulse Flow not mA Input. The user must override the factory default and select mA Input to enable the current input functionality. Upon selecting mA Input, the following menu screen will appear to allow complete programming of mA Current Input. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Configure Measure: mA Input mA Input: Temperature Units: ºC Input Range: 4­20mA Low Value: High Value : Filter: 0.001% 100.0% 5sec To program current input, scroll to the desired item and press ENTER. The following sub­sections provide you with the initial display screen that appears for each programming routine. Use the flow diagram for current input programming at the end of Sec. 6 and the live screen prompts to complete programming. 6.11.2 Measurement The display screen for selecting the signal board func­ tionality is shown. The default value is displayed in bold type. Scroll down to select mA Input to enable the current input functionality. Refer to the current input Programming flow diagram to complete this func­ tion. 6.11.3 mA Input The display screen for selecting the type of measure­ ment is shown. The default measurement type for mA Input is displayed in bold type. Refer to the current input Programming flow diagram to complete this function. 64 S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Measurement Pulse Flow mA Input S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN mA Input Temperature Pressure Flow Other MODEL 1056 SECTION 6.0 PROGRAMMING THE MEASUREMENTS 6.11.4 Units The display screen for selecting measurement units is shown. The default value for temperature is displayed in bold type. Refer to the current input Programming flow diagram to complete this function. If Pressure is selected as the measurement type for mA Input, the following display screen is shown: S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Units °C ºF S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Units mm Hg in Hg atm kPa mbar bar The current input board can also be used to accept a 4­20mA current input from a pulse flow sensor. If Flow is selected as the measurement type for the 4­20mA current input board, the following display screen is shown: S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Units GPM GPH cu ft/min cu ft/hour L/min L/hour m3/hour Current input can serve as a universal measurement board. 4­20mA current input can be accepted from any device and assigned to represent a wide range of measurements. If Other is selected as the measure­ ment type for the 4­20mA current input board, the fol­ lowing display screen is shown: S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Units % % Saturation pH mV Any of the following units can also be selected to represent the 4­20mA current input. Simply scroll down to identify and select the desired measurement units as listed in the table below. µS/cm ppm µg/L NTU ft/sec mS/cm ppb m/sec mg/L FTU MΩ­cm g/L FNU kΩ­cm ‰ 6.11.5 Input Range The display screen for selecting the Input Range is shown. The default value for mA Input is displayed in bold type. Refer to the current input Programming flow diagram to complete this function. none S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Input Range 4­20mA 0­20mA 65 MODEL 1056 6.11.6 Low Value The display screen for entering the Low Value to be assigned to 4mA (or 0mA) current input is shown. The default value for temperature is displayed in bold type. Refer to the current input Programming flow diagram to complete this function. 6.11.7 High Value The display screen for entering the High Value to be assigned to 20mA current input is shown. The default value for temperature is displayed in bold type. Refer to the current input Programming flow diagram to com­ plete this function. 6.11.8 Filter The display screen for entering the input filter value in seconds is shown. The default value is displayed in bold type. Refer to the current input Programming diagram to complete this function. . 66 SECTION 6.0 PROGRAMMING THE MEASUREMENTS S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Low Value 0.000ºC S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN High Value 100.0ºC S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Input Filter 005sec SECTION 6.0 PROGRAMMING THE MEASUREMENTS FIGURE 6­1 Configuring pH/ORP Measurements MODEL 1056 67 FIGURE 6­2 Configure Contacting Measurements MODEL 1056 68 SECTION 6.0 PROGRAMMING THE MEASUREMENTS SECTION 6.0 PROGRAMMING THE MEASUREMENTS FIGURE 6­3 Configure Toroidal Measurements MODEL 1056 69 FIGURE 6­5 Configure Oxygen Measurements MODEL 1056 70 SECTION 6.0 PROGRAMMING THE MEASUREMENTS MODEL 1056 SECTION 6.0 PROGRAMMING THE MEASUREMENTS FIGURE 6­4 Configure Chlorine Measurements FIGURE 6­6 Configure Ozone Measurements 71 SECTION 6.0 PROGRAMMING THE MEASUREMENTS FIGURE 6­7 Configure Turbidity Measurement MODEL 1056 72 MODEL 1056 SECTION 6.0 PROGRAMMING THE MEASUREMENTS FIGURE 6­8 Configure Flow Measurement FIGURE 6­9 Configure mA Current Input Measurement 73 MODEL 1056 74 SECTION 6.0 PROGRAMMING THE MEASUREMENTS This page left blank intentionally MODEL 1056 SECTION 7.0 CALIBRATION SECTION 7.0 CALIBRATION 7.1 7.2 7.3 7.4 7.5 7.6 CALIBRATION – INTRODUCTION pH CALIBRATION ORP CALIBRATION CONTACTING CONDUCTIVITY CALIBRATION TOROIDAL CONDUCTIVITY CALIBRATION CHLORINE CALIBRATION 7.6.1 FREE CHLORINE 7.6.2 TOTAL CHLORINE 7.6.3 MONOCHLORAMINE 7.6.4 pH­INDEPENDENT FREE CHLORINE 7.7 OXYGEN CALIBRATION 7.8 OZONE CALIBRATION 7.9 TEMPERATURE CALIBRATION 7.10 TURBIDITY CALIBRATION 7.11 FLOW CALIBRATION 7.1 CALIBRATION – INTRODUCTION Calibration is the process of adjusting or standardizing the analyzer to a lab test or a calibrated laboratory instrument, or standardizing to some known reference (such as a commercial buffer). The auto­recognition feature of the analyzer will enable the appropriate calibration screens to allow calibration for any single sensor configuration or dual sensor config­ uration of the analyzer. Completion of Quick Start upon first power up enables live measurements but does not ensure accurate readings in the lab or in process. Calibration should be performed with each attached sensor to ensure accurate, repeatable readings. This section covers the following programming and configuration functions: 4. Standardization calibration (1­point) for pH, ORP and Redox (pH Cal ­ Sec.7.2 and 7.3) 5. Entering the cell constant of a conductivity sensor (Conductivity Cal ­ Sec. 7.4 and 7.5) 6. Calibrating the sensor in a conductivity standard Conductivity Cal ­ Sec. 7.4 and 7.5) 7. Calibrating the analyzer to a laboratory instrument (Contacting Conductivity Cal ­ Sec.7.4) 8. Zeroing an chlorine, oxygen or ozone sensor (Amperometric Cal ­ Sec’s 7.6, 7.7, 7.8) 9. Calibrating an oxygen sensor in air (Oxygen Cal ­ Sec’s 7.6) 10. Calibrating the sensor to a sample of known concentration (Amperometric Cal ­ Sec’s 7.6, 7.7, 7.8) 11. Enter a manual reference temperature for temperature compensation of the process measurement 1. Auto buffer cal for pH (pH Cal ­ Sec.7.2) 2. Manual buffer cal for pH (pH Cal ­ Sec.7.2) 3. Set calibration stabilization criteria for pH (pH Cal ­ Sec.7.2) 75 MODEL 1056 SECTION 7.0 CALIBRATION 7.2 pH CALIBRATION 7.2.1 DESCRIPTION New sensors must be calibrated before use. Regular recalibration is also necessary. Use auto calibration instead of manual calibration. Auto calibration avoids common pitfalls and reduces errors. The analyzer recognizes the buffers and uses temperature­corrected pH values in the calibration. Once the Model 1056 successfully completes the calibration, it calcu­ lates and displays the calibration slope and offset. The slope is reported as the slope at 25°C. THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1056 WITH A pH SENSOR. THE FOLLOWING CALIBRATION ROUTINES ARE COVERED. TABLE 7­1 Measure Sec. pH 7.2.2 pH Calibration Routines Menu function: default Description Auto Calibration ­ pH 2 point buffer calibration with auto buffer recognition 7.2.3 Manual Calibration ­ pH 2 point buffer calibration with manual buffer value entry 7.2.4 Entering A Known Slope Value ­ pH Slope calibration with manual entry of known slope value 7.2.5 Standardization ­ 1 point buffer calibration with manual buffer value entry pH A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines. To calibrate pH: 1. Press the MENU button 2. Select Calibrate. Press ENTER. 3. Select Sensor 1 or Sensor 2 corresponding to pH. Press ENTER. 4. Select pH. Press ENTER. The following screen will appear. To calibrate pH or Temperature scroll to the desired item and press ENTER. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibrate? pH Temperature The following sub­sections show the initial display screen that appears for each calibration routine. Use the flow diagram for pH calibration at the end of Sec. 7 and the live screen prompts to complete calibration. 7.2.2 AUTO CALIBRATION — pH This screen appears after selecting pH calibration. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN pH Cal Buffer Cal Standardize Slope: 59.16mV/pH Offset: 600 mV Note that pH auto calibration criteria can be changed. The following criteria can be adjusted:  Stabilization time (default 10 sec.)  Stabilization pH value (default 0.02 pH)  Type of Buffer used for AUTO CALIBRATION (default is Standard, non­commercial buffers). The following commercial buffer tables are recognized by the analyzer:  Standard (NIST plus pH7)  DIN 19267  Ingold  Merck 76 The following screen will appear to allow adjustment of these criteria: S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Setup Stable Time: 10 sec Stable Delta: 0.02 pH Buffer: Standard MODEL 1056 SECTION 7.0 CALIBRATION The following screen will appear if the auto cal is successful. The screen will return to the pH Buffer Cal Menu. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN pH Auto Cal Slope: 59.16 mV/pH Offset: 60 mV The following screens may appear if the auto cal is unsuccessful. 1. A High Slope Error will generate this screen display: S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN pH Auto Cal High Slope Error Calculated: 62.11 mV/pH Max: 62.00 mV/pH Press EXIT 2. A Low Slope Error will generate this screen display: S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN pH Auto Cal Low Slope Error Calculated: 39.11mV/pH Min: 40.00 mV/pH Press EXIT 3. An Offset Error will generate this screen display: S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN pH Auto Cal Offset Error Calculated: 61.22mV Max: 60.00mV Press EXIT 7.2.3 MANUAL CALIBRATION — pH New sensors must be calibrated before use. Regular recalibration is also necessary. Use manual calibration if non­standard buffers are being used; otherwise, use auto calibration. Auto calibration avoids common pitfalls and reduces errors. The adjacent appears after selecting Manual pH calibra­ 7.2.4 ENTERING A KNOWN SLOPE VALUE — pH If the electrode slope is known from other measure­ ments, it can be entered directly in the Model 1056 ana­ lyzer. The slope must be entered as the slope at 25°C. 7.2.5 STANDARDIZATION — pH The pH measured by the Model 1056 analyzer can be changed to match the reading from a second or referee instrument. The process of making the two readings agree is called standardization. During standardization, the difference between the two pH values is converted to the equivalent voltage. The voltage, called the reference offset, is added to all subsequent measured cell voltages before they are converted to pH. If a standardized sensor is placed in a buffer solution, the measured pH will differ from the buffer pH by an amount equivalent to the standardization offset. tion. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN pH Manual Cal Buffer 1 Buffer 2 S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN pH Slope@25ºC 59.16 mV/pH S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Enter Value 07.00pH 77 MODEL 1056 SECTION 7.0 CALIBRATION The following screen may appear if ORP Cal is unsuccessful. An Offset Error will generate this screen display: S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Standardize Offset Error Calculated: 96mV Max: 60mV Press EXIT If the ORP Cal is successful, the screen will return to the Cal sub­menu. 7.3 ORP CALIBRATION 7.3.1 DESCRIPTION For process control, it is often important to make the measured ORP agree with the ORP of a standard solution. During calibration, the measured ORP is made equal to the ORP of a standard solution at a single point. THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1056 WITH AN ORP SENSOR. THE FOL­ LOWING CALIBRATION ROUTINE IS COVERED. TABLE 7­2 Measure ORP ORP Calibration Routine Sec. 7.3.2 Menu function: default Description Standardization — ORP 1 point buffer calibration with manual buffer value entry A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines. To calibrate ORP: 1. Press the MENU button 2. Select Calibrate. Press ENTER. 3. Select Sensor 1 or Sensor 2 corresponding to ORP. Press ENTER. 4. Select ORP. Press ENTER. The following screen will appear. To calibrate ORP or Temperature, scroll to the desired item and press ENTER. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibrate? ORP Temperature The following sub­sections show the initial display screen that appears for each calibration routine. Use the flow diagram for ORP calibration at the end of Sec. 7 and the live screen prompts to complete calibration. 7.3.2 STANDARDIZATION — ORP For process control, it is often important to make the measured ORP agree with the ORP of a standard solution. During calibration, the measured ORP is made equal to the ORP of a standard solution at a single point. This screen appears after selecting ORP cali­ bration: Cal sub­menu. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Enter Value +0600 mV If the ORP Cal is successful, the screen will return to the The following screen may appear if ORP Cal is unsuccessful. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Standardize Offset Error Calculated: 61.22mV Max: 60.00mV Press EXIT 78 MODEL 1056 SECTION 7.0 CALIBRATION 7.4 CONTACTING CONDUCTIVITY CALIBRATION 7.4.1 DESCRIPTION PLACING A NEW CONDUCTIVITY SENSOR IN SERVICE New conductivity sensors rarely need calibration. The cell constant printed on the label is sufficiently accurate for most applications. CALIBRATING AN IN­SERVICE CONDUCTIVITY SENSOR 1. After a conductivity sensor has been in service for a period of time, recalibration may be necessary. There are three ways to calibrate a sensor. a. Use a standard instrument and sensor to measure the conductivity of the process stream. It is not necessary to remove the sensor from the process piping. The temperature correction used by the standard instrument may not exactly match the temperature correction used by the Model 1056. To avoid errors, turn off temperature correction in both the analyzer and the standard instrument. b. Place the sensor in a solution of known conductivity and make the analyzer reading match the conductivity of the standard solution. Use this method if the sensor can be easily removed from the process piping and a standard is available. Be careful using standard solutions having conductivity less than 100 µS/cm. Low conductivity standards are highly susceptible to atmospheric contamination. Avoid calibrating sensors with 0.01/cm cell constants against conductivity standards having conductivity greater than 100 µS/cm. The resistance of these solutions may be too low for an accurate measurement. Calibrate sensors with 0.01/cm cell constant using method c. c. To calibrate a 0.01/cm sensor, check it against a standard instrument and 0.01/cm sensor while both sensors are measuring water having a conductivity between 5 and 10 µS/cm. To avoid drift caused by absorption of atmospheric carbon dioxide, saturate the sample with air before making the measurements. To ensure adequate flow past the sensor during calibration, take the sample downstream from the sensor. For best results, use a flow­through standard cell. If the process temperature is much different from ambient, keep connecting lines short and insulate the flow cell. THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1056 WITH AN ATTACHED CONTACTING CONDUCTIVITY SENSOR. THE FOLLOWING CALIBRATION ROUTINES ARE COVERED. TABLE 7­3 Contacting Conductivity Calibration Routines Menu function: default Description Measure Sec. Contacting Conductivity 7.4.2 Cell K: 7.4.3 Zero Cal Zero the analyzer with the sensor attached 7.4.4 In Process Cal Standardize the sensor to a known conductivity 7.4.5 Meter Cal Calibrate the analyzer to a lab conductivity instrument 7.4.6 Cal Factor: 1.00000/cm 0.95000/cm Enter the cell Constant for the sensor Enter the Cal Factor for 4­Electrode sensors from the sensor tag A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines To calibrate contacting conductivity: 1. Press the MENU button 2. Select Calibrate. Press ENTER. S1: 1.234µS/cm 25.0ºC 3. Select Sensor 1 or Sensor 2 corresponding to S2: 12.34pH 25.0ºC contacting conductivity. Press ENTER. SN Calibrate? 4. Select Conductivity. Press ENTER. The adjacent screen will appear. To calibrate Conductivity or Temperature, scroll to the desired item and press ENTER. The following sub­sections show the initial display screen that appears for each calibration routine. Use the flow diagram for Conductivity calibration at the end of Sec. 7 and the live screen prompts for each rou­ tine to complete calibration. The adjacent screen appears Conductivity calibration: after selecting Conductivity Temperature S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibration Zero Cal In Process Cal Meter Cal Cell K: 1.00000/cm 79 MODEL 1056 7.4.2 ENTERING THE CELL CONSTANT New conductivity sensors rarely need calibration. The cell constant printed on the label is sufficiently accurate for most applications. The cell constant should be entered: • When the unit is installed for the first time • When the probe is replaced The display screen for entering a cell Constant for the sensor is shown. The default value is displayed in bold type. 7.4.3 ZEROING THE INSTRUMENT This procedure is used to compensate for small offsets to the conductivity signal that are present even when there is no conductivity to be measured. This procedure is affected by the length of extension cable and should always be repeated if any changes in extension cable or sensor have been made. Electrically connect the conductivity probe as it will actually be used and place the measuring portion of the probe in air. Be sure the probe is dry. The adjacent screen will appear after selecting Zero Cal from the Conductivity Calibration screen: The adjacent screen will appear if zero Cal is successful. The screen will return to the conductivity Cal Menu. SECTION 7.0 CALIBRATION S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Cell Constant 1.00000 /cm S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Zero Cal In Air In Water S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Zero Cal Sensor Zero Done The adjacent screen may appear if zero Cal is unsuccessful. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Zero Cal Sensor Zero Fail Offset too high Press EXIT 7.4.4 CALIBRATING THE SENSOR IN A CONDUCTIVITY STANDARD (IN PROCESS CAL) This procedure is used to calibrate the sensor and analyzer against a solution of known conductivity. This is done by submerging the probe in the sample of known conductivity, then adjusting the displayed value, if necessary, to correspond to the conductivity value of the sample. Turn temperature correction off and use the conductivity of the standard. Use a calibrated ther­ mometer to measure temperature. The probe must be cleaned before performing this procedure. The adjacent screen will appear after selecting In Process Cal from the Conductivity Calibration screen: 80 S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN InProcess Cal Wait for stable reading. MODEL 1056 SECTION 7.0 CALIBRATION The adjacent screen will appear if In Process Cal is suc­ cessful. The screen will return to the conductivity Cal Menu. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC The adjacent screen may appear if In Process Cal is unsuccessful. The screen will return to the conductivity Cal Men S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN InProcess Cal Updated cell constant: 1.00135/cm SN InProcess Cal Calibration Error Press EXIT 7.4.5 CALIBRATING THE SENSOR TO A LABORATORY INSTRUMENT (METER CAL) This procedure is used to check and correct the conductivity reading of the Model 1056 using a laboratory conductivity instrument. This is done by submerging the conductivity probe in a bath and measuring the conduc­ tivity of a grab sample of the same bath water with a separate laboratory instrument. The Model 1056 reading is then adjusted to match the conductivity reading of the lab instrument. The adjacent screen will appear after selecting Meter Cal from the Conductivity Calibration screen: After pressing ENTER, the display shows the live value measured by the sensor S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Meter Cal Use precision resistors only S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Enter Value xx.xx kΩ If the meter cal is successful the screen will return to the conductivity Cal Menu. The adjacent screen will appear if Meter Cal is unsuccessful. The screen will return to the conductivity Cal Menu. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Meter Cal Calibration Error Press EXIT 7.4.6 Cal Factor Upon initial installation and power up, if 4­electrode was selected for the sensor type in the Quick Start menus, the user enters a Cell Constant and a “Cal Factor” using the instrument keypad. The cell constant is needed to convert measured conductance to conduc­ tivity as displayed on the analyzer screen. The “Cal Factor” entry is needed increase the accuracy of the live conductivity readings, especially at low conductivity readings below 20uS/cm. Both the Cell Constant and the “Cal Factor” are printed on the tag attached to the 4­electrode sensor/cable. The display screen for entering Cal Factor is shown. The default value is displayed in bold type. If neces­ sary after initial installation and start­up, enter the “Cal Factor” as printed on the sensor tag. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Cal Factor 0.95000 /cm 81 MODEL 1056 SECTION 7.0 CALIBRATION 7.5 TOROIDAL CONDUCTIVITY CALIBRATION 7.5.1 DESCRIPTION Calibration is the process of adjusting or standardizing the analyzer to a lab test or a calibrated laboratory instru­ ment, or standardizing to some known reference (such as a conductivity standard). This section contains procedures for the first time use and for routine calibration of the Model 1056 analyzer. THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1056 WITH AN ATTACHED INDUC­ TIVE/TOROIDAL CONDUCTIVITY SENSOR. THE FOLLOWING CALIBRATION ROUTINES ARE COVERED TABLE 7­4 Measure Toroidal Conductivity Toroidal Conductivity Calibration Sec. Calibration function: default value Description 7.5.2 Cell K: 3.00000/cm Enter the cell Constant for the sensor 7.5.3 Zero Cal Zeroing the analyzer with the sensor attached 7.5.4 In Process Cal Standardizing the sensor to a known conductivity A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines. To calibrate toroidal conductivity: 1. Press the MENU button 2. Select Calibrate. Press ENTER. 3. Select Sensor 1 or Sensor 2 corresponding to Toroidal Conductivity. Press ENTER. 4. Select Conductivity. Press ENTER. The adjacent screen will appear. To calibrate Toroidal Conductivity or Temperature, scroll to the desired item and press ENTER The following sub­sections show the initial display screen that appears for each calibration routine. Use the flow diagram for Conductivity calibration at the end of Sec. 7 and the live screen prompts to complete calibration. The adjacent screen appears after selecting Conductivity calibration: 82 S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibrate? Conductivity Temperature S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibration Zero Cal In Process Cal Cell K: 1.00000/cm MODEL 1056 SECTION 7.0 CALIBRATION 7.5.2 ENTERING THE CELL CONSTANT New conductivity sensors rarely need calibration. The cell constant printed on the label is sufficiently accurate for most applications. The cell constant should be entered: • When the unit is installed for the first time • When the probe is replaced • During troubleshooting This procedure sets up the analyzer for the probe type connected to the analyzer. Each type of probe has a specific cell constant: S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Cell Constant 3.00000 /cm The display screen for entering a cell constant for the sensor is shown. The default value is displayed in bold type. 7.5.3 ZEROING THE INSTRUMENT This procedure is used to compensate for small offsets to the conductivity signal that are present even when there is no conductivity to be measured. This procedure is affected by the length of extension cable and should always be repeated if any changes in extension cable or sensor have been made. Electrically connect the conductivity probe as it will actually be used and place the measuring portion of the probe in air. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Zero Cal In Air In Water The adjacent screen will appear after selecting Zero Cal from the Conductivity Calibration screen: The adjacent screen will appear if zero Cal is successful. The screen will return to the conductivity Cal Menu. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Zero Cal Sensor Zero Done . The adjacent screen may appear if zero Cal is unsuccessful. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Zero Cal Sensor Zero Fail Offset too high Press EXIT 7.5.4 CALIBRATING THE SENSOR IN A CONDUCTIVITY STANDARD (IN PROCESS CAL) This procedure is used to check and correct the conductivity reading of the Model 1056 to ensure that the reading is accurate. This is done by submerging the probe in the sample of known conductivity, then adjusting the displayed value, if necessary, to correspond to the conductivity value of the sample. The probe must be cleaned before performing this procedure. The temper­ ature reading must also be checked and standardized if necessary, prior to performing this procedure. The adjacent screen will appear after selecting In Process Cal from the Conductivity Calibration screen: S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN InProcess Cal Wait for stable reading. 83 MODEL 1056 The following screen will appear if In Process Cal is successful. The screen will return to the conductivity Cal Menu. This screen may appear if In Process Cal is unsuccessful. The screen will return to the conductivity Cal Menu. SECTION 7.0 CALIBRATION S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN InProcess Cal Updated cell constant: 3.01350/cm S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN InProcess Cal Calibration Error Press EXIT 7.6 CALIBRATION —CHLORINE With a Chlorine measurement board and the appropriate sensor, Model 1056 can measure any of four variants of Chlorine: • Free Chlorine • Total Chlorine • Monochloramine • pH­independent Free Chlorine The section describes how to calibrate any compatible amperometric chlorine sensor. The following calibration routines are covered in the family of supported Chlorine sensors: • Air Cal • Zero Cal • In Process Cal 7.6.1 CALIBRATION — FREE CHLORINE 7.6.1.1 DESCRIPTION A free chlorine sensor generates a current directly proportional to the concentration of free chlorine in the sam­ ple. Calibrating the sensor requires exposing it to a solution containing no chlorine (zero standard) and to a solu­ tion containing a known amount of chlorine (full­scale standard). The zero calibration is necessary because chlo­ rine sensors, even when no chlorine is in the sample, generate a small current called the residual current. The analyzer compensates for the residual current by subtracting it from the measured current before converting the result to a chlorine value. New sensors require zeroing before being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. Either of the following makes a good zero standard: • Deionized water containing about 500 ppm sodium chloride. Dissolve 0.5 grams (1/8 teaspoonful) of table salt in 1 liter of water. DO NOT USE DEIONIZED WATER ALONE FOR ZEROING THE SENSOR. THE CONDUCTIVITY OF THE ZERO WATER MUST BE GREATER THAN 50 μS/cm. • Tap water known to contain no chlorine. Expose tap water to bright sunlight for at least 24 hours. The purpose of the In Process calibration is to establish the slope of the calibration curve. Because stable chlorine standards do not exist, the sensor must be calibrated against a test run on a grab sample of the process liquid. Several manufacturers offer portable test kits for this purpose. Observe the following precautions when taking and testing the grab sample. • Take the grab sample from a point as close to the sensor as possible. Be sure that taking the sample does not alter the flow of the sample to the sensor. It is best to install the sample tap just downstream from the sensor. • Chlorine solutions are unstable. Run the test immediately after taking the sample. Try to calibrate the sensor when the chlorine concentration is at the upper end of the normal operating range. THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1056 WITH A FREE CHLORINE SENSOR. THE FOLLOWING CALIBRATION ROUTINES ARE COVERED. 84 MODEL 1056 TABLE 7­5 Measure Free Chlorine SECTION 7.0 CALIBRATION Free Chlorine Calibration Routines Sec. Calibration function: default value Description 7.6.1.2 Zero Cal Zeroing the sensor in solution with zero free chlorine 7.6.1.3 In Process Cal Standardizing to a sample of known chlorine concentration A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines. To calibrate free chlorine: 1. Press the MENU button 2. Select Calibrate. Press ENTER. 3. Select Sensor 1 or Sensor 2 corresponding to Free Chlorine. Press ENTER. 4. Select Free Chlorine. Press ENTER. The adjacent screen will appear. To calibrate Free Chlorine or Temperature, scroll to the desired item and press ENTER. The following sub­sections show the initial display screen that appears for each calibration routine. Use the flow diagram for Chlorine calibration at the end of Sec. 7 and the live screen prompts to complete cal­ ibration. The adjacent screen appears after selecting Free Chlorine calibration: 7.6.1.2 ZEROING THE SENSOR. The adjacent screen will appear during Zero Cal. Be sure sensor has been running in zero solution for at least two hours before starting zero step. The adjacent screen will appear if In Zero Cal is successful. The screen will return to the Amperometric Cal Menu. The adjacent screen may appear if In Zero Cal is unsuc­ cessful. The screen will return to the Amperometric Cal Menu. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibrate? Free Chlorine Temperature S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibration Zero Cal In Process Cal S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Zero Cal Zeroing Wait S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Zero Cal Sensor zero done S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Zero Cal Sensor zero failed Press EXIT 7.6.1.3 IN PROCESS CALIBRATION The adjacent screen will appear prior to In Process Cal S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN InProcess Cal Wait for stable reading. 85 MODEL 1056 SECTION 7.0 CALIBRATION If the In Process Cal is successful, the screen will return to the Cal sub­menu. The adjacent screen may appear if In Zero Cal is unsuc­ cessful. The screen will return to the Amperometric Cal Menu. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN InProcess Cal Calibration Error Press EXIT 7.6.2 CALIBRATION — TOTAL CHLORINE 7.6.2.1 DESCRIPTION Total chlorine is the sum of free and combined chlorine. The continuous determination of total chlorine requires two steps. First, the sample flows into a conditioning system (TCL) where a pump continuously adds acetic acid and potassium iodide to the sample. The acid lowers the pH, which allows total chlorine in the sample to quantitatively oxidize the iodide in the reagent to iodine. In the second step, the treated sample flows to the sensor. The sensor is a membrane­covered amperometric sensor, whose output is proportional to the concentration of iodine. Because the concentration of iodine is proportional to the concentration of total chlorine, the analyzer can be calibrated to read total chlorine. Because the sensor really measures iodine, calibrating the sensor requires exposing it to a solution containing no iodine (zero standard) and to a solution containing a known amount of iodine (full­scale standard). The Zero calibration is necessary because the sensor, even when no iodine is present, generates a small current called the residual current. The analyzer compensates for the residual current by subtracting it from the measured current before converting the result to a total chlorine value. New sensors require zeroing before being placed in service, and sensors should be TABLE 7­ 6 Measure Total Chlorine zeroed whenever the electrolyte solution is replaced. The best zero standard is deionized water. The purpose of the In Process Calibration is to establish the slope of the calibration curve. Because stable total chlorine standards do not exist, the sensor must be calibrated against a test run on a grab sample of the process liquid. Several manufac­ turers offer portable test kits for this purpose. Observe the following precautions when taking and testing the grab sample: • Take the grab sample from a point as close as possible to the inlet of the TCL sample conditioning system. Be sure that taking the sample does not alter the flow through the TCL. • Chlorine solutions are unstable. Run the test immedi­ ately after taking the sample. Try to calibrate the sensor when the chlorine concentration is at the upper end of the normal operating range. Note this measurement must be made using the Model TCL total chlorine sample conditioning system. THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1056 WITH AN ATTACHED TOTAL CHLORINE SENSOR. THE FOLLOWING CALIBRATION ROUTINES ARE COVERED. Total Chlorine Calibration Routines Sec. Calibration function: default value Description 7.6.2.2 Zero Cal Zeroing the sensor in solution with zero total chlorine 7.6.2.3 In Process Cal Standardizing to a sample of known chlorine concentration A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines. To calibrate total chlorine: 1. Press the MENU button 2. Select Calibrate. Press ENTER. 3. Select Sensor 1 or Sensor 2 corresponding to Total Chlorine. Press ENTER. 4. Select Total Chlorine. Press ENTER. The adjacent screen will appear. To calibrate Total Chlorine or Temperature, scroll to the desired item and press ENTER 86 S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibrate? Total Chlorine Temperature MODEL 1056 The following sub­sections provide you with the initial display screen that appears for each calibration routine. Use the flow diagram for Chlorine calibration at the end of Sec. 7 and the live screen prompts to complete calibration. SECTION 7.0 CALIBRATION S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibration Zero Cal In Process Cal This adjacent screen appears after selecting Total Chlorine calibration: 7.6.2.2 ZEROING THE SENSOR. The adjacent screen will appear during Zero Cal. Be sure sensor has been running in zero solution for at least two hours before starting zero step. The adjacent screen will appear if In Zero Cal is suc­ cessful. The screen will return to the Amperometric Cal Menu. The adjacent screen may appear if In Zero Cal is unsuccessful. The screen will return to the Amperometric Cal Menu. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Zero Cal Zeroing Wait S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Zero Cal Sensor zero done S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Zero Cal Sensor zero failed Press EXIT 7.6.2.3 IN PROCESS CALIBRATION The adjacent screen will appear prior to In Process Cal S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN InProcess Cal Wait for stable reading. If the In Process Cal is successful, the screen will return to the Cal sub­menu. The adjacent screen may appear if In Process Cal is unsuccessful. The screen will return to the Amperometric Cal Menu. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN InProcess Cal Calibration error Press EXIT 87 MODEL 1056 SECTION 7.0 CALIBRATION 7.6.3 CALIBRATION ­ MONOCHLORAMINE 7.6.3.1 DESCRIPTION A monochloramine sensor generates a current directly proportional to the concentration of monochloramine in the sample. Calibrating the sensor requires exposing it to a solution containing no monochloramine (zero standard) and to a solution containing a known amount of monochloramine (full­scale standard). The Zero calibration is necessary because monochloramine sensors, even when no monochloramine is in the sample, generate a small current called the residual or zero current. The analyzer compensates for the residual current by subtracting it from the measured current before converting the result to a monochloramine value. New sensors require zeroing before being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. The best zero standard is deionized water. The purpose of the In Process calibration is to establish the slope of the calibration curve. Because stable monochloramine standards do not exist, the sensor must be calibrated against a test run on a grab sample of the process liquid. Several manufacturers offer portable test kits for this purpose. Observe the following precautions when taking and testing the grab sample. • Take the grab sample from a point as close to the sen­ sor as possible. Be sure that taking the sample does not alter the flow of the sample to the sensor. It is best to install the sample tap just downstream from the sensor. • Monochloramine solutions are moderately unstable. Run the test as soon as possible after taking the sam­ ple. Try to calibrate the sensor when the monochlo­ ramine concentration is at the upper end of the normal operating range. THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1056 WITH AN ATTACHED MONO­ CHLORAMINE SENSOR. THE FOLLOWING CALIBRATION ROUTINES ARE COVERED. TABLE 7­7 Monochloramine Calibration Routines Measure Monochloramine 7.6.3.2 Zero Cal Zeroing the sensor in solution with zero monochloramine 7.6.3.3 In Process Cal Standardizing to a sample of known chlorine concentration Sec. Calibration function: default value Description A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines. To calibrate monochloramine: 1. Press the MENU button 2. Select Calibrate. Press ENTER. 3. Select Sensor 1 or Sensor 2 corresponding to Monochloramine. Press ENTER. 4. Select Monochloramine. Press ENTER. The adjacent screen will appear. To calibrate Monochloramine or Temperature, scroll to the desired item and press ENTER. The following sub­sections provide you with the initial display screen that appears for each calibration routine. Use the flow diagram for Chlorine calibration at the end of Sec. 7 and the live screen prompts to complete calibration. The adjacent screen appears Monochloramine calibration: 88 after selecting S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibrate? Monochloramine Temperature S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibration Zero Cal In Process Cal MODEL 1056 7.6.3.2 ZEROING THE SENSOR. The adjacent screen will appear during Zero Cal. Be sure sensor has been running in zero solution for at least two hours before starting zero step. The adjacent screen will appear if In Zero Cal is suc­ cessful. The screen will return to the Amperometric Cal Menu. The adjacent screen may appear if In Zero Cal is unsuccessful. The screen will return to the Amperometric Cal Menu. SECTION 7.0 CALIBRATION S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Zero Cal Zeroing Wait S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Zero Cal Sensor zero done S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Zero Cal Sensor zero failed Press EXIT 7.6.3.3 IN PROCESS CALIBRATION The adjacent screen will appear prior to In Process Cal S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN InProcess Cal Wait for stable reading. If the In Process Cal is successful, the screen will return to the Cal sub­menu. The adjacent screen may appear if In Process Cal is unsuccessful. The screen will return to the Amperometric Cal Menu. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN InProcess Cal Calibration Error Press EXIT 89 MODEL 1056 SECTION 7.0 CALIBRATION 7.6.4 pH­INDEPENDENT FREE CHLORINE MEASUREMENT 7.6.4.1 DESCRIPTION A free chlorine sensor generates a current directly proportional to the concentration of free chlorine in the sample. Calibrating the sensor requires exposing it to a solution containing no chlorine (zero standard) and to a solution containing a known amount of chlorine (full­ scale standard). The zero calibration is necessary because chlorine sensors, even when no chlorine is in the sample, generate a small current called the residual current. The analyzer compensates for the residual current by subtracting it from the measured current before converting the result to a chlorine value. New sensors require zeroing before being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. Either of the following makes a good zero standard: • Deionized water. • Tap water known to contain no chlorine. Expose tap water to bright sunlight for at least 24 hours. The purpose of the In Process calibration is to establish the slope of the calibration curve. Because stable chlorine standards do not exist, the sensor must be calibrated against a test run on a grab sample of the process liquid. Several manufacturers offer portable test kits for this purpose. Observe the following precautions when taking and testing the grab sample. • Take the grab sample from a point as close to the sensor as possible. Be sure that taking the sample does not alter the flow of the sample to the sensor. It is best to install the sample tap just downstream from the sensor. • Chlorine solutions are unstable. Run the test immediately after taking the sample. Try to calibrate the sensor when the chlorine concentration is at the upper end of the normal operating range. Note: This measurement is made using the model 498CL­01 ­ pH­independent Free Chlorine sensor manufactured by Rosemount Analytical. THIS SECTION DESCRIBES HOW TO CALIBRATE T H E M O D E L 1 0 5 6 W I T H A N AT TA C H E D PH­ INDEPENDENT FREE CHLORINE SENSOR. THE FOLLOWING CALIBRATION ROUTINES ARE COVERED. TABLE 7­ 8 Measure pH­independent Free Chlorine pH­independent Free Chlorine Calibration Routines Sec. Calibration function: default value Description 7.6.4.2 Zero Cal Zeroing the sensor in solution with zero free chlorine 7.6.4.3 In Process Cal Standardizing to a sample of known chlorine concentration A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines. To calibrate pH­independent free chlorine: 1. Press the MENU button 2. Select Calibrate. Press ENTER. 3. Select Sensor 1 or Sensor 2 corresponding to pH­independent free chlorine. Press ENTER. 4. Select pH Ind. Free Cl. Press ENTER. The adjacent screen will appear. To calibrate pH­independent Free Chlorine or Temperature, scroll to the desired item and press ENTER. 90 S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibrate? pH Ind. Free Cl Temperature MODEL 1056 The following sub­sections provide you with the initial display screen that appears for each calibration routine. Use the flow diagram for Chlorine calibration at the end of Sec. 7 and the live screen prompts to complete calibration. The adjacent screen appears after selecting pH­inde­ pendent free chlorine calibration: 7.6.4.2 ZEROING THE SENSOR. The adjacent screen will appear during Zero Cal SECTION 7.0 CALIBRATION S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibration Zero Cal In Process Cal S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Zero Cal Zeroing Wait The adjacent screen will appear if In Zero Cal is suc­ cessful. The screen will return to the Amperometric Cal Menu. The adjacent screen may appear if In Zero Cal is unsuccessful. The screen will return to the Amperometric Cal Menu. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Zero Cal Sensor zero done S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Zero Cal Sensor zero failed Press EXIT 7.6.4.3 IN PROCESS CALIBRATION The following screen will appear prior to In Process Cal S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN InProcess Cal Wait for stable reading. If the In Process Cal is successful, the screen will return to the Cal sub­menu. The adjacent screen may appear if In Process Cal is unsuccessful. The screen will return to the Amperometric Cal Menu. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN InProcess Cal Calibration Error Press EXIT 91 MODEL 1056 7.7 CALIBRATION — OXYGEN 7.7.1 DESCRIPTION Oxygen sensors generate a current directly proportional to the concentration of dissolved oxygen in the sample. Calibrating the sensor requires exposing it to a solution containing no oxygen (zero standard) and to a solution containing a known amount of oxygen (full­scale standard). The Zero Calibration is necessary because oxygen sensors, even when no oxygen is present in the sample, generate a small current called the residual current. The analyzer compensates for the residual current by subtracting it from the measured current before converting the result to a dissolved oxygen value. New sensors require zeroing before being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. The recommended zero standard is 5% sodium sulfite in water, although oxygen­free nitrogen can also be used. The Model 499A TrDO sensor, used for the determination of trace (ppb) oxygen levels, has very low residual current and does not normally require zeroing. The residual current in the 499A TrDO sensor is equivalent to less than 0.5 ppb oxygen. The purpose of the In Process Calibration is to establish the slope of the calibration curve. Because the solubility of atmospheric oxygen in water as a function of temperature and barometric pressure is well known, the natural choice for a full­scale standard is air­saturated water. However, air­saturated water is difficult to prepare and use, so the universal practice is to use air for calibration. From the point of view of the oxygen sensor, air and air­saturated water are identical. The equivalence comes about because the sensor really measures the chemical potential of oxygen. Chemical potential is the force that causes oxygen molecules to diffuse from the sample into the sensor where they can be measured. It is also the force that causes oxygen molecules in air to dissolve in water and to continue to dissolve until the water is saturated with oxygen. Once the water is saturated, the chemical potential of oxygen in the two phases (air and water) is the same. Oxygen sensors generate a current directly proportional to the rate at which oxygen molecules diffuse through a membrane stretched over the end of the sensor. The 92 SECTION 7.0 CALIBRATION diffusion rate depends on the difference in chemical potential between oxygen in the sensor and oxygen in the sample. An electrochemical reaction, which destroys any oxygen molecules entering the sensor, keeps the concentration (and the chemical potential) of oxygen inside the sensor equal to zero. Therefore, the chemical potential of oxygen in the sample alone deter­ mines the diffusion rate and the sensor current. When the sensor is calibrated, the chemical potential of oxygen in the standard determines the sensor current. Whether the sensor is calibrated in air or air­saturated water is immaterial. The chemical potential of oxygen is the same in either phase. Normally, to make the calculation of solubility in common units (like ppm DO) simpler, it is convenient to use water­saturated air for calibration. Automatic air calibration is standard. The user simply exposes the sensor to water­saturated air. The analyzer monitors the sensor current. When the current is stable, the analyzer stores the current and measures the temperature using a temperature element inside the oxygen sensor. The user must enter the barometric pressure. From the temperature the analyzer calculates the saturation vapor pressure of water. Next, it calculates the pressure of dry air by subtracting the vapor pressure from the barometric pressure. Using the fact that dry air always contains 20.95% oxygen, the analyzer calcu­ lates the partial pressure of oxygen. Once the analyzer knows the partial pressure of oxygen, it uses the Bunsen coefficient to calculate the equilibrium solubility of atmospheric oxygen in water at the prevailing tem­ perature. At 25°C and 760 mm Hg, the equilibrium sol­ ubility is 8.24 ppm. Often it is too difficult or messy to remove the sensor from the process liquid for calibra­ tion. In this case, the sensor can be calibrated against a measurement made with a portable laboratory instru­ ment. The laboratory instrument typically uses a mem­ brane­covered amperometric sensor that has been cali­ brated against water­saturated air. THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1056 WITH AN OXYGEN SENSOR. THE FOLLOWING CALIBRATION ROUTINES ARE COVERED. MODEL 1056 TABLE 7­ 9 Measure Oxygen SECTION 7.0 CALIBRATION Oxygen Calibration Routines Sec. Calibration function: default value Description 7.7.2 Zero Cal Zeroing the sensor in a medium with zero oxygen 7.7.3 Air Cal Calibrating the sensor in a water­saturated air sample 7.7.4 In Process Cal Standardizing to a sample of known oxygen concentration 7.7.5 Sen@ 25°C:2500nA/ppm Entering a known slope value for sensor response 7.7.6 Zero Current: 0nA Entering a known zero current for a specific sensor A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines. Oxygen sensors generate a current directly proportional to the concentration of dissolved oxygen in the sample. Calibrating the sensor requires exposing it to a solution containing no oxygen (zero standard) and to a solution containing a known amount of oxygen (full­scale standard). Automatic air calibration is standard. The user simply exposes the sensor to water­saturated air. To calibrate oxygen: 1. Press the MENU button 2. Select Calibrate. Press ENTER. 3. Select Sensor 1 or Sensor 2 corresponding to oxygen. Press ENTER. 4. Select Oxygen. Press ENTER. The adjacent screen will appear. To calibrate Oxygen or Temperature, scroll to the desired item and press ENTER The following sub­sections provide you with the initial display screen that appears for each calibration routine. Use the flow diagram for Oxygen calibration at the end of Sec. 7 and the live screen prompts for each routine to complete calibration. The adjacent screen appears after selecting Oxygen calibration: Air calibration criteria can be changed. The following criteria can be adjusted:  Stabilization time (default 10 sec.)  Stabilization pH value (default 0.05 ppm)  Salinity of the solution to be measured (default 00.0 parts per thousand) The adjacent screen will appear to allow adjustment of these criteria” S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibrate? Oxygen Temperature S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibration Air Cal Zero Cal In Process Cal Sen@ 25°C:2500nA/ppm Zero Current: 1234nA S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Setup Stable Time: 10 sec Stable Delta: 0.05 ppm Salinity: 00.0 ‰ 93 MODEL 1056 SECTION 7.0 CALIBRATION 7.7.2 ZEROING THE SENSOR. The adjacent screen will appear during Zero Cal . S1: 1.234 nA S2: 1.456 nA SN Zero Cal Zeroing Wait The adjacent screen will appear if In Zero Cal is suc­ cessful. The screen will return to the Amperometric Cal Menu. S1: 1.234 nA S2: 1.456 nA The adjacent screen may appear if In Zero Cal is unsuccessful. The screen will return to the Amperometric Cal Menu. S1: 1.234 nA S2: 1.456 nA SN Zero Cal Sensor zero done SN Zero Cal Sensor zero failed Press EXIT 7.7.3 CALIBRATING THE SENSOR IN AIR The adjacent screen will appear prior to Air Cal S1: 1.234µS/cm S2: 12.34pH 25.0ºC 25.0ºC SN Air Cal Start Calibration Setup The adjacent screen will appear if In Air Cal is suc­ cessful. The screen will return to the Amperometric Cal Menu. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC The adjacent screen will appear if In Air Cal is suc­ cessful. The screen will return to the Amperometric Cal Menu. S1: 1.234µS/cm S2: 12.34pH SN Air Cal Done 25.0ºC 25.0ºC SN Air Cal Failure Check Sensor Press EXIT 7.7.4 CALIBRATING THE SENSOR AGAINST A STANDARD INSTRUMENT (IN PROCESS CAL) S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC The adjacent screen will appear prior to In Process Cal SN InProcess Cal Wait for stable reading. If the In Process Cal is successful, the screen will return to the Cal sub­menu. S1: 1.234µS/cm S2: 12.34pH The adjacent screen may appear if In Zero Cal is unsuccessful. The screen will return to the Amperometric Cal Menu. 94 25.0ºC 25.0ºC SN InProcess Cal Calibration Error Press EXIT MODEL 1056 SECTION 7.0 CALIBRATION 7.8 CALIBRATION — OZONE 7.8.1 DESCRIPTION An ozone sensor generates a current directly proportional to the concentration of ozone in the sample. Calibrating the sensor requires exposing it to a solution containing no ozone (zero standard) and to a solution containing a known amount of ozone (full­scale standard). The Zero Calibration is necessary because ozone sensors, even when no ozone is in the sample, generate a small current called the residual or zero current. The analyzer compensates for the residual current by subtracting it from the measured current before converting the result to an ozone value. New sensors require zeroing before being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. The best zero standard is deionized water. against a test run on a grab sample of the process liquid. Several manufacturers offer portable test kits for this purpose. Observe the following precautions when taking and testing the grab sample. • Take the grab sample from a point as close to the sensor as possible. Be sure that taking the sample does not alter the flow of the sample to the sensor. It is best to install the sample tap just downstream from the sensor. • Ozone solutions are unstable. Run the test immedi­ ately after taking the sample. Try to calibrate the sensor when the ozone concentration is at the upper end of the normal operating range. THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1056 WITH AN OZONE SENSOR. THE FOLLOWING CALIBRATION ROUTINES ARE COV­ ERED. The purpose of the In Process Calibration is to establish the slope of the calibration curve. Because stable ozone standards do not exist, the sensor must be calibrated TABLE 7­ 10 Measure Ozone Ozone Calibration Routines Sec. Calibration function: default value Description 7.8.2 Zero Cal Zeroing the sensor in solution with zero ozone 7.8.3 In Process Cal Standardizing to a sample of known ozone concentration A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines. To calibrate ozone: 1. Press the MENU button 2. Select Calibrate. Press ENTER. 3. Select Sensor 1 or Sensor 2 corresponding to ozone. Press ENTER. 4. Select Ozone. Press ENTER. The adjacent screen will appear. To calibrate Ozone or Temperature, scroll to the desired item and press ENTER. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibrate? Ozone Temperature 95 MODEL 1056 The following sub­sections provide you with the initial display screen that appears for each calibration routine. Use the flow diagram for Ozone calibration at the end of Sec. 7 and the live screen prompts to complete calibration. SECTION 7.0 CALIBRATION S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibration Zero Cal In Process Cal The adjacent screen appears after selecting Ozone calibration: 7.8.2 ZEROING THE SENSOR. The following screen will appear during Zero Cal S1: S2: The following screen will appear if In Zero Cal is successful. The screen will return to the Amperometric Cal Menu. S1: 1.234 nA S2: 1.456 nA The following screen may appear if In Zero Cal is unsuccessful. The screen will return to the Amperometric Cal Menu. S1: S2: 1.234 nA 1.456 nA SN Zero Cal Zeroing Wait SN Zero Cal Sensor zero done 1.234 nA 1.456 nA SN Zero Cal Sensor zero failed Press EXIT 7.8.3 IN PROCESS CALIBRATION The following screen will appear after selecting In Process Cal S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC If the In Process Cal is successful, the screen will return to the Cal sub­menu. S1: 1.234µS/cm S2: 12.34pH The following screen may appear if In Zero Cal is unsuccessful. The screen will return to the Amperometric Cal Menu. 96 SN InProcess Cal Wait for stable reading. 25.0ºC 25.0ºC SN InProcess Cal Calibration Error Press EXIT MODEL 1056 SECTION 7.0 CALIBRATION 7.9 CALIBRATING TEMPERATURE 7.9.1 DESCRIPTION Most liquid analytical measurements require temperature compensation (except ORP). The Model 1056 performs temperature compensation automatically by applying internal temperature correction algorithms. Temperature cor­ rection can also be turned off. If temperature correction is off, the Model 1056 uses the manual temperature entered by the user in all temperature correction calculations. THIS SECTION DESCRIBES HOW TO CALIBRATE TEMPERATURE IN THE MODEL 1056 ANALYZER. THE FOLLOWING CALIBRATION ROUTINE IS COVERED. TABLE 7­ 11 Measure Temperature Temperature Calibration Routine Sec. 7.9.2 Calibration function: default value Calibrate Description Enter a manual reference temperature for temperature compensation of the process measurement A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines. To calibrate temperature: 1. Press the MENU button 2. Select Calibrate. Press ENTER. 3. Select Sensor 1 or Sensor 2 corresponding to the desired measurement. Press ENTER. 4. Select Temperature. Press ENTER. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibrate +025.0°C The adjacent screen will appear. The following sub­section provides you with the initial display screen that appears for temperature calibration. Use the flow diagram for Temp calibration at the end of Sec. 7 to complete calibration. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibrate Cal in progress. Please wait. 7.9.2 CALIBRATION The adjacent screen will appear during Temperature Cal. If the sensor Temperature offset is greater than 5 ºC from the default value, the following screen will appear: S1: 1.234µS/cm S2: 12.34pH You may continue by selecting Yes or suspend this operation by selecting No. No Yes 25.0ºC 25.0ºC SN Temp Offset > 5°C Continue? If the Temp Cal is successful, the screen will return to the Cal Menu. Note: To select automatic or manual temp compensation or to program temperature units as °C or °F, refer to Sec. 5.3 – Programming Temperature in this manual 97 MODEL 1056 SECTION 7.0 CALIBRATION 7.10 TURBIDITY 7.10.1 DESCRIPTION This section describes how to calibrate the turbidity sensor against a user­prepared standard as a 2­point calibra­ tion with di­ionized water, against a 20 NTU user­prepared standard as a single point calibration, and against a grab sample using a reference turbidimeter. THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1056 WITH AN ATTACHED TURBIDITY SENSOR AS PART OF THE COMPLETE CLARITY II TURBIDITY SYSTEM. THE FOLLOWING CALIBRATION ROUTINES ARE COVERED. TABLE 7­12 TURBIDITY CALIBRATION ROUTINES Measure Sec. Calibration function: Turbidity 7.10.2 Slope Calibration default value Description Slope cal with pure water and a standard of known turbidity 7.10.3 Standardize Calibration Standardizing the sensor to a known turbidity 7.10.4 Grab Calibration Standardizing the sensor to a known turbidity based on a reference turbidimeter A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines. To calibrate Turbidity: 1. Press the MENU button 2. Select Calibrate. Press ENTER. 3. Select Sensor 1 or Sensor 2 corresponding to Turbidity. Press ENTER. 4. Select Turbidity. Press ENTER. The following sub­sections provide you with the initial display screen that appears for each calibration rou­ tine. Use the flow diagram for Turbidity calibration at the end of Sec. 7 and the live screen prompts to complete calibration. 7.10.2 SLOPE CALIBRATION — Turbidity This section describes how to conduct a 2­point cali­ bration of the turbidity sensor against a user­prepared 20NTU standard. The calibration requires two steps. First, immerse the sensor in filtered water having very low turbidity and measure the sensor output. Next, increase the turbidity of the filtered water by a known amount, typically 20 NTU, and measure the sensor output again. The analyzer takes the two measure­ ments, applies a linearization correction (if necessary), and calculates the sensitivity. Sensitivity is the sensor output (in mV) divided by turbidity. A typical new sen­ sor has a sensitivity of about 10 mV/NTU. As the sen­ sor ages, the sensitivity decreases. The figure below illustrates how turbidity calibration works. Before beginning the calibration, the analyzer does a dark current measurement. Dark current is the signal gener­ ated by the detector when no light is falling on it. The analyzer subtracts the dark current from the raw scat­ 98 The following screen will appear. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibrate? Turbidity S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibrate Slope Standard Grab tered light signal and converts the result to turbidity. In highly filtered samples, which scatter little light, the dark current can be a substantial amount of the signal generated by the detector. MODEL 1056 This screen appears after selecting Slope calibration. SECTION 7.0 CALIBRATION S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Slope Cal Sensor in pure H2O? Press ENTER The following screen will appear if Slope Cal is suc­ cessful. The screen will return to the Turbidity Cal Menu. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC The following screen may appear if Slope Cal is unsuccessful. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Slope Cal Cal Complete SN Slope Cal Calibration Error Press EXIT 7.10.3 STANDARDIZE CALIBRATION ­ Turbidity The turbidity sensor can also be calibrated against a commercial standard. Stable 20.0 NTU standards are available from a number of sources. Calibration using a commercial standard is simple. Filtered deionized water is not required. Before beginning the calibration, the analyzer does a dark current measurement. Dark cur­ rent is the signal generated by the detector even when no light is falling on it. The analyzer subtracts the dark current from the raw scattered light signal and converts the result to turbidity. In highly filtered samples, which scatter little light, the dark current can be a substantial amount of the signal generated by the sensor. This screen appears after selecting Standard calibration. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Standard Cal Sensor in Standard? Press ENTER The following screen will appear if Standard Cal is successful. The screen will return to the Turbidity Cal Menu. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC The following screen may appear if Standard Cal is unsuccessful. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Standard Cal Cal Complete SN Standard Cal Calibration Error Press EXIT 99 MODEL 1056 SECTION 7.0 CALIBRATION 7.10.4 GRAB CALIBRATION ­ Turbidity If desired, the turbidity sensor can be calibrated against the turbidity reading from another instrument. The analyzer treats the value entered by the user as though it were the true turbidity of the sample. Therefore, grab sample calibration changes the sensi­ tivity, it does not apply an offset to the reading. This screen appears after selecting Grab calibration. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Grab Cal Wait for stable reading The following screen will appear if Grab Cal is successful. The screen will return to the Turbidity Cal Menu. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Grab Cal Cal Complete The following screen may appear if Grab Cal is unsuccessful. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Grab Cal Calibration Error Press EXIT 7.11 PULSE FLOW 7.11.1 DESCRIPTION A variety of pulse flow sensors can be wired to the Flow signal input board to measure flow volume, total volume and flow difference (if 2 Flow signal boards are installed). The Model 1056 Flow signal board will support flow sensors that are self­driven (powered by the rotation of the impeller paddle­wheel). THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1056 WITH AN ATTACHED FLOW SENSOR. THE FOLLOWING CALIBRATION ROUTINES ARE COVERED. TABLE 7­13 FLOW CALIBRATION ROUTINES Measure Pulse Flow Calibration function Description 7.11.2 K Factor A constant value representing pulses/Gal of flow 7.11.3 Frequency/Velocity & Pipe Alternate cal method – requires manual entry of frequency (Hz) per velocity and Pipe diameter used 7.11.4 In process Calibration Calibration based on known volume per unit of time 7.11.5 Totalizer Control User settings to stop, restart and reset total volume meter Sec. A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines. To 1. 2. 3. calibrate Pulse Flow: Press the MENU button Select Calibrate. Press ENTER. Select Sensor 1 or Sensor 2 corresponding to Flow. Press ENTER. 4. Select Pulse Flow. Press ENTER. The following screen will appear. 100 S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibrate? Pulse Flow MODEL 1056 To calibrate Pulse Flow scroll to the desired item and press ENTER. The following sub­sections provide you with the initial display screen that appears for each calibration rou­ tine. Use the diagram for Pulse Flow calibration at the end of Sec. 7 and the live screen prompts to com­ plete calibration. 7.11.2 CALIBRATION — K Factor This screen appears after selecting K Factor. SECTION 7.0 CALIBRATION . S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibration K Factor: 12.34 p/Gal Freq/Velocity & Pipe In Process Totalizer Control S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN K Factor 12.34 p/Gal . To calibrate Pulse Flow scroll to the desired item and press ENTER. The following sub­sections provide you with the initial display screen that appears for each calibration rou­ tine. Use the diagram for Pulse Flow calibration at the end of Sec. 7 and the live screen prompts to com­ plete calibration. . S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Calibration K Factor: 12.34 p/Gal Freq/Velocity & Pipe In Process Totalizer Control Simply enter the know K factor provided with the flow sensor specifications. The screen will return to the Pulse Flow Cal Menu and the updated K factor will appear. 7.11.3 CALIBRATION — Freq/Velocity & Pipe This screen appears after selecting Freq/Velocity & Pipe calibration. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC After completing the entry of the Freq/Velocity ratio, the following screen will appear: S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC The following screen will appear if the entries are suc­ cessful. The screen will return to the Pulse Flow Cal Menu. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Freq/Velocity 12.34 Hz per ft/sec SN Pipe Diameter 10.00 in SN Freq/Velocity&Pipe Updated K Factor 12.34 p/Gal 101 MODEL 1056 The following screen may appear if the entries are unsuccessful. SECTION 7.0 CALIBRATION S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Freq/Velocity&Pipe Calibration Error Press EXIT 7.11.4 CALIBRATION — In Process Cal This screen appears after selecting In Process Cal. The following screen will appear if the entries are suc­ cessful. The screen will return to the Pulse Flow Cal Menu. The following screen will appear if the entries are suc­ cessful. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN InProcess Cal Updated K Factor 12.34 p/Gal S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN InProcess Cal Calibration Error Press EXIT 7.11.5 CALIBRATION — Totalizer Control This screen appears after selecting Totalizer Control. S1: 1.234µS/cm 25.0ºC S2: 12.34pH 25.0ºC SN Totalizer Control Stop Resume Reset 123456789012.3 G The user can suspend the totalizer by selecting Stop, re­enable the totalizer by selecting Resume and return the totalizer volume count to zero by selecting Reset. The live totalizer volume count is displayed during these menu operations. 102 SECTION 7.0 CALIBRATION FIGURE 7­1 Calibrate pH MODEL 1056 103 FIGURE 7­2 Calibrate ORP MODEL 1056 104 SECTION 7.0 CALIBRATION FIGURE 7­3 Calibrate Contacting and Toroidal Conductivity MODEL 1056 SECTION 7.0 CALIBRATION 105 FIGURE 7­4 Calibrate Free Chlorine, Total Chlorine, Monochloramine, and pH­independent Free Chlorine MODEL 1056 106 SECTION 7.0 CALIBRATION SECTION 7.0 CALIBRATION FIGURE 7­5 Calibrate Oxygen MODEL 1056 107 FIGURE 7­6 Calibrate Ozone MODEL 1056 108 SECTION 7.0 CALIBRATION SECTION 7.0 CALIBRATION FIGURE 7­7 Calibrate Temperature MODEL 1056 109 FIGURE 7­8 Calibrate Turbidity MODEL 1056 110 SECTION 7.0 CALIBRATION SECTION 7.0 CALIBRATION FIGURE 7­9 Calibrate Flow MODEL 1056 111 MODEL 1056 SECTION 8.0 RETURN OF MATERIAL SECTION 8.0 RETURN OF MATERIAL 17.1 GENERAL. To expedite the repair and return of instruments, proper communication between the customer and the factory is important. Before returning a product for repair, call 1­949­757­8500 for a Return Materials Authorization (RMA) number. 17.2 WARRANTY REPAIR. The following is the procedure for returning instruments still under warranty: 1. Call Rosemount Analytical for authorization. 2. To verify warranty, supply the factory sales order number or the original purchase order number. In the case of individual parts or sub­assemblies, the serial number on the unit must be supplied. 3. Carefully package the materials and enclose your “Letter of Transmittal” (see Warranty). If possible, pack the materials in the same manner as they were received. 4. Send the package prepaid to: Emerson Process Management Rosemount Analytical 2400 Barranca Parkway Irvine, CA 92606 Attn: Factory Repair RMA No. ____________ IMPORTANT Please see second section of “Return of Materials Request” form. Compliance with the OSHA requirements is mandatory for the safety of all personnel. MSDS forms and a certification that the instruments have been disinfected or detoxified are required. Mark the package: Returned for Repair Model No. ____ 17.3 NON­WARRANTY REPAIR. The following is the procedure for returning for repair instruments that are no longer under warranty: 1. Call Rosemount Analytical for authorization. 2. Supply the purchase order number, and make sure to provide the name and telephone number of the individ­ ual to be contacted should additional information be needed. 3. Do Steps 3 and 4 of Section 17.2. NOTE Consult the factory for additional information regarding service or repair. 112 The right people, the right answers, right now. Immediate, Reliable Analytical Support Now there’s a way to quickly get the right answers for your liquid analytical instrumentation questions: the Analytical Customer Support Center. Our staff of trained professionals is ready to provide the information you need. If you are placing an order, verifying delivery, requesting application information, or just want to contact a Rosemount Analytical representative, a call to the Customer Support Center will provide you with the right people, the right answers, right now. A Worldwide Network of Sales and Service Emerson Process Management’s field sales offices are your source for more information on the fill line of Rosemount Analytical products. Field sales personnel will work closely with you to supply technical data and application information. For more information, please contact your nearest Emerson Process Management sales office. THE AMERICAS ­ HEADQUARTERS Emerson Process Management Rosemount Analytical Inc. Liquid Center of Excellence 2400 Barranca Parkway Irvine, CA 92606 Phone: +1.949.757.8500 Toll Free: +1.800.854.8257 Fax: +1.949.474.7250 ASIA­PACIFIC Emerson Process Management Asia Pacific Private Ltd. 1 Pandan Crescent Singapore 0512 Republic of Singapore Phone: 65.777.8211 Fax: 65.777.0947 GERMANY Emerson Process Management Process Gas Analyzer Center of Excellence GmbH & Co. OHG Industriestrasse 1 63594 Hasselroth Germany T 49.6055.884.0 F 49.6055.884.20 LATIN AMERICA Emerson Process Management Rosemount Analytical 10241 West Little York, Suite #200 Houston, TX 77040 USA T 713.467.6000 F 713.827.3328 MIDDLE EAST AND AFRICA EUROPE Emerson Process Management AG Blegistrasse 21 CH­6341 Baar­Walterswil Switzerland T 41.41.768.6111 T 41.41.761.8740 VISIT OUR WEBSITE AT www.rosemountanalytical.com Emerson Process Management EPM Building P. O. Box 17033 Jebe Ali Free Zone Dubai, United Arab Emirates T 971.4.8835235 F 971.4.8835312 The right people, the right answers, right now. ON­LINE ORDERING NOW AVAILABLE ON OUR WEB SITE http://www.rosemountanalytical.com 8 Credit Cards for U.S. Purchases Only. Emerson Process Management 2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757­8500 Fax: (949) 474­7250 http://www.rosemountanalytical.com © Rosemount Analytical Inc. 2014