Emerson 1056 DUAL-INPUT INTELLIGENT ANALYZER Instruction Manual
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Instruction Manual
PN 511056/rev.J
May 2014
DUAL-INPUT INTELLIGENT ANALYZER
1056
ESSENTIAL INSTRUCTIONS
WARNING
READ THIS PAGE BEFORE PROCEEDING!
Your instrument purchase from Rosemount
Analytical, Inc. is one of the finest available for your
particular application. These instruments have been
designed, and tested to meet many national and
international standards. Experience indicates that its
performance is directly related to the quality of the
installation and knowledge of the user in operating
and maintaining the instrument. To ensure their con
tinued operation to the design specifications, per
sonnel should read this manual thoroughly before
proceeding with installation, commissioning, opera
tion, and maintenance of this instrument. If this
equipment is used in a manner not specified by the
manufacturer, the protection provided by it against
hazards may be impaired.
• Failure to follow the proper instructions may
cause any one of the following situations to
occur: Loss of life; personal injury; property dam
age; damage to this instrument; and warranty
invalidation.
• Ensure that you have received the correct model
and options from your purchase order. Verify that
this manual covers your model and options. If
not, call 18008548257 or 9497578500 to
request correct manual.
• For clarification of instructions, contact your
Rosemount representative.
RISK OF ELECTRICAL SHOCK
Equipment protected throughout by double insulation.
• Installation and servicing of this product may expose personel
to dangerous voltages.
• Main power wired to separate power source must be
disconnected before servicing.
• Do not operate or energize instrument with case open!
• Signal wiring connected in this box must be rated at least
240 V.
• Nonmetallic cable strain reliefs do not provide grounding
between conduit connections! Use grounding type bushings
and jumper wires.
• Unused cable conduit entries must be securely sealed by
nonflammable closures to provide enclosure integrity in
compliance with personal safety and environmental protection
requirements. Unused conduit openings must be sealed with
NEMA 4X or IP65 conduit plugs to maintain the ingress
protection rating (NEMA 4X).
• Electrical installation must be in accordance with the National
Electrical Code (ANSI/NFPA70) and/or any other applicable
national or local codes.
• Operate only with front panel fastened and in place.
• Safety and performance require that this instrument be
connected and properly grounded through a threewire
power source.
• Proper use and configuration is the responsibility of the
user.
• Follow all warnings, cautions, and instructions
marked on and supplied with the product.
• Use only qualified personnel to install, operate,
update, program and maintain the product.
• Educate your personnel in the proper installation,
operation, and maintenance of the product.
• Install equipment as specified in the Installation
section of this manual. Follow appropriate local
and national codes. Only connect the product to
electrical and pressure sources specified in this
manual.
• Use only factory documented components for
repair. Tampering or unauthorized substitution of
parts and procedures can affect the performance
and cause unsafe operation of your process.
• All equipment doors must be closed and protec
tive covers must be in place unless qualified per
sonnel are performing maintenance.
Emerson Process Management
2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 7578500
Fax: (949) 4747250
http://www.rosemountanalytical.com
© Rosemount Analytical Inc. 2012
CAUTION
This product generates, uses, and can radiate radio frequency
energy and thus can cause radio communication interference.
Improper installation, or operation, may increase such interfer
ence. As temporarily permitted by regulation, this unit has not
been tested for compliance within the limits of Class A comput
ing devices, pursuant to Subpart J of Part 15, of FCC Rules,
which are designed to provide reasonable protection against
such interference. Operation of this equipment in a residential
area may cause interference, in which case the user at his own
expense, will be required to take whatever measures may be
required to correct the interference.
CAUTION
This product is not intended for use in the light industrial,
residential or commercial environments per the instru
ment’s certification to EN500812.
QUICK START GUIDE
1056 Dual Input Analyzer
1. Refer to Section 2.0 for mechanical installation instructions.
2.
Wire sensor(s) to the signal boards. See Section 3.0 for wiring instructions. Refer to the sensor instruction
sheet for additional details. Make current output, alarm relay and power connections.
3. Once connections are secured and verified, apply power to the analyzer.
WARNING
RISK OF ELECTRICAL SHOCK
Electrical installation must be in accordance with
the National Electrical Code (ANSI/NFPA70)
and/or any other applicable national or local codes.
4. When the analyzer is powered up for the first time, Quick Start screens appear. Quick Start operating tips
are as follows:
a. A backlit field shows the position of the cursor.
b. To move the cursor left or right, use the keys to the left or right of the ENTER key. To scroll up or down
or to increase or decrease the value of a digit use the keys above and below the ENTER key . Use the
left or right keys to move the decimal point.
c. Press ENTER to store a setting. Press EXIT to leave without storing changes. Pressing EXIT during Quick
Start returns the display to the initial startup screen (select language).
5. Complete the steps as shown in the Quick Start Guide flow diagram, Fig. A on the following page.
6. After the last step, the main display appears. The outputs are assigned to default values.
7.
To change output, and temperaturerelated settings, go to the main menu and choose Program. Follow the
prompts. For a general guide to the Program menu, see the Quick Reference Guide, Fig.B.
8. To return the analyzer to the default settings, choose Reset Analyzer under the Program menu.
Figure A. QUICK START GUIDE
QUICK START GUIDE
Figure B. MODEL 1056 MENU TREE
QUICK REFERENCE GUIDE
About This Document
This manual contains instructions for installation and operation of the Model 1056 DualInput Intelligent
Analyzer. The following list provides notes concerning all revisions of this document.
Rev. Level
Date
Notes
A
01/07
B
C
2/07
9/07
D
11/07
E
F
G
H
I
J
05/08
08/08
09/08
04/10
03/12
05/14
This is the initial release of the product manual. The manual has been reformatted to reflect the
Emerson documentation style and updated to reflect any changes in the product offering.
Added CE mark to p.2. Replaced Quick Start Fig A.
Revised Sections 1,3,5,6, and 7. Added new measurements and features Turbidity, Flow, Current
Input, Alarm relays and 4electrode conductivity.
Added 24VDC power supply to Sec. 3.4. Added CSA and FM agency approvals for option codes
01, 20, 21, 22, 24, 25, 26, 30, 31, 32, 34, 35, 36 and 38.
Add HART and Profibus DP digital communication to Section 1 specifications.
Updates
FM and CSA agency approval, Class 1, Div 2. for 24 VDC and AC switching power supplies.
Update DNV logo and company name
Update addresses mail and web
Update enclosure information
MODEL 1056
TABLE OF CONTENTS
MODEL 1056
DUAL INPUT INTELLIGENT ANALYZER
TABLE OF CONTENTS
QUICK START GUIDE
QUICK REFERENCE GUIDE
TABLE OF CONTENTS
Section Title
1.0
DESCRIPTION AND SPECIFICATIONS ................................................................
2.0
INSTALLATION .......................................................................................................
2.1
Unpacking and Inspection........................................................................................
2.2
Installation................................................................................................................
Page
1
11
11
11
3.0
3.1
3.2
3.3
3.4
WIRING....................................................................................................................
General ....................................................................................................................
Preparing Conduit Openings....................................................................................
Preparing Sensor Cable ..........................................................................................
Power, Output, Alarms and Sensor Connections.....................................................
19
19
19
20
20
4.0
4.1
4.2
4.3
4.4
5.0
DISPLAY AND OPERATION ...................................................................................
User Interface ..........................................................................................................
Instrument Keypad...................................................................................................
Main Display ............................................................................................................
Menu System ...........................................................................................................
PROGRAMMING – BASICS ...................................................................................
27
27
27
28
29
31
5.1
5.2
5.3
5.4
5.5
5.6
5.7
5.8
5.9
6.0
General ....................................................................................................................
Changing the StartUp Settings ................................................................................
Choosing Temperature units and Automatic/Manual Temperature Compensation ..
Configuring and Ranging the Current Outputs.........................................................
Setting a Security Code ...........................................................................................
Security Access........................................................................................................
Using Hold ...............................................................................................................
Resetting Factory Defaults – Reset Analyzer ..........................................................
Alarm Relays............................................................................................................
PROGRAMMING MEASUREMENTS ...................................................................
31
31
32
32
34
35
35
36
37
41
6.1
6.2
6.3
6.4
6.5
6.6
Programming Measurements – Introduction ...........................................................
pH ............................................................................................................................
ORP .........................................................................................................................
Contacting Conductivity ..........................................................................................
Toroidal Conductivity................................................................................................
Chlorine....................................................................................................................
41
42
43
45
48
51
6.6.1 Free Chlorine ..................................................................................................
51
6.6.2 Total Chlorine .................................................................................................
53
6.6.3 Monochloramine ............................................................................................
54
6.6.4 pHindependent Free Chlorine .......................................................................
55
Oxygen.....................................................................................................................
Ozone ......................................................................................................................
57
59
6.7
6.8
i
MODEL 1056
TABLE OF CONTENTS
TABLE OF CONTENTS CONT’D
6.9
Turbidity ...................................................................................................................
60
6.10
Flow .........................................................................................................................
63
6.11
Current Input ............................................................................................................
64
7.0
7.1
CALIBRATION ......................................................................................................
Calibration – Introduction .........................................................................................
75
75
7.2
pH Calibration ..........................................................................................................
76
7.3
ORP Calibration .......................................................................................................
78
7.4
Contacting Conductivity Calibration .........................................................................
79
7.5
Toroidal Conductivity Calibration .............................................................................
82
7.6
Chlorine Calibration .................................................................................................
84
7.6.1 Free Chlorine ..................................................................................................
84
7.6.2 Total Chlorine ..................................................................................................
86
7.6.3 Monochloramine .............................................................................................
88
7.6.4 pHIndependent Free Chlorine .......................................................................
90
7.7
Oxygen Calibration ..................................................................................................
92
7.8
Ozone Calibration ....................................................................................................
95
7.9
Temperature Calibration...........................................................................................
97
7.10
Turbidity ...................................................................................................................
98
7.11
Pulse Flow ...............................................................................................................
100
8.0
RETURN OF MATERIAL ........................................................................................
Warranty...................................................................................................................
112
112
ii
MODEL 1056
TABLE OF CONTENTS
TABLE OF CONTENTS CONT’D
LIST OF FIGURES
Fig#
A
B
21
22
23
24
25
26
31
32
33
34
35
36
37
38
39
310
311
312
313
314
315
41
51
52
53
54
55
61
62
63
64
65
66
67
68
69
71
72
73
74
75
76
77
78
79
Section
PREFACE
PREFACE
SEC 2.0
SEC 2.0
SEC 2.0
SEC 2.0
SEC 2.0
SEC 2.0
SEC 3.4
SEC 3.4
SEC 3.4
SEC 3.4
SEC 3.4
SEC 3.4
SEC 3.4
SEC 3.4
SEC 3.4
SEC 3.4
SEC 3.4
SEC 3.4
SEC 3.4
SEC 3.4
SEC 3.4
SEC 4.3
SEC 5.3.2
SEC 5.4.5
SEC 5.5.2
SEC 5.7.2
SEC 5.8.2
SEC 6.2
SEC 6.4
SEC 6.5
SEC 6.6
SEC 6.7
SEC 6.8
SEC 6.9
SEC 6.10
SEC 6.11
SEC 7.2
SEC 7.3
SEC 7.4
SEC 7.6
SEC 7.7
SEC 7.8
SEC 7.9
SEC 7.10
SEC 7.11
Figure Title
Quick Start Guide
Quick Reference Guide
Panel Mounting Dimensions ..............................................................
Pipe and Wall Mounting Dimensions .................................................
CSA Certification drawing part 1 ........................................................
CSA Certification drawing part 2 ........................................................
FM NonIncendive drawing part 1......................................................
FM NonIncendive drawing part 2......................................................
115/230 VAC Power Supply ...............................................................
24VDC Power Supply ........................................................................
Switching AC Power Supply...............................................................
Current Output Wiring ........................................................................
Alarm Relay Wiring for Model 1056 Switching Power Supply............
Contacting Conductivity board and sensor cable leads .....................
Toroidal Conductivity signal board and sensor cable leads ...............
pH/ORP/ISE signal board and sensor cable leads ............................
Amperometric board (Cl, O2, Ozone) and sensor cable leads ..........
Turbidity signal board wiht plugin Sensor connection.......................
Flow/Current Input signal board and Sensor cable leads ..................
Power Wiring for Model 1056 115/230 VAC.......................................
Power Wiring for Model 1056 85265 VAC ........................................
Output Wiring for Model 1056 Main PCB...........................................
Power Wiring for Model 1056 24VDC ................................................
Formatting the Main Display .............................................................
Choosing Temp Units and Manual Auto Temp Compensation ...........
Configuring and Ranging the Current Outputs...................................
Setting A Security Code ....................................................................
Using Hold .........................................................................................
Resetting Factory Default Settings ....................................................
Configuring pH/ORP Measurements .................................................
Configure Contacting Measurements ...............................................
Configure Toroidal Measurements ....................................................
Configure Chlorine Measurements ....................................................
Configure Oxygen Measurements .....................................................
Configure Ozone Measurements .......................................................
Configure Turbidity Measurement......................................................
Configure Flow Measurement............................................................
Configure mA Current Input Measurement ........................................
Calibrate pH .......................................................................................
Calibrate ORP....................................................................................
Calibrate Contacting and Toroidal Conductivity .................................
Calibrate Chlorine ..............................................................................
Calibrate Oxygen ...............................................................................
Calibrate Ozone .................................................................................
Calibrate Temperature .......................................................................
Calibrate Turbidity ..............................................................................
Calibrate Flow ....................................................................................
Page
12
13
14
15
16
17
20
20
20
21
21
22
22
23
23
24
24
25
25
26
26
30
32
33
34
35
36
67
68
69
71
70
71
72
73
73
103
104
105
106
107
108
109
110
111
iii
MODEL 1056
TABLE OF CONTENTS
TABLE OF CONTENTS CONT’D
LIST OF TABLES
Number Section
iv
Table Title
Page
51
SEC 5.2.1
Measurements and Measurement Units .............................................
31
61
SEC 6.2.1
pH Measurement Programming ...........................................................
42
62
SEC 6.3.1
ORP Measurement Programming ........................................................
43
63
SEC 6.4.1
Contacting Conductivity Measurement Programming..........................
45
64
SEC 6.5.1
Toroidal Conductivity Measurement Programming...............................
48
65
SEC 6.6.1.1 Free Chlorine Measurement Programming..........................................
51
66
SEC 6.6.2.1 Total Chlorine Measurement Programming .........................................
53
67
68
69
610
611
612
613
71
72
73
74
75
76
77
78
79
710
711
712
713
SEC 6.6.3.1Monochloramine Measurement Programming.....................................
SEC 6.6.4 pHIndependent Free Chlorine Measurement Programming ............
SEC. 6.7.1 Oxygen Measurement Programming....................................................
SEC 6.8.1 Ozone Measurement Programming......................................................
SEC 6.9.1 Turbidity Measurement Programming..................................................
SEC 6.10.1 Flow Measurement Programming.........................................................
SEC 6.11.1 Curent Input Programming...................................................................
SEC 7.2
pH Calibration Routines .......................................................................
SEC 7.3
ORP Calibration Routine......................................................................
SEC 7.4
Contacting Conductivity Calibration Routines......................................
SEC 7.5
Toroidal Conductivity Calibration ........................................... ............
SEC 7.6.1 Free Chlorine Calibration Routines.......................................................
SEC 7.6.2 Total Chlorine Calibration Routines......................................................
SEC 7.6.3 Monochloramine Calibration Routines..................................................
SEC 7.6.4 pHindependent Free Chlorine Calibration Routines............................
SEC 7.7
Oxygen Calibration Routines................................................................
SEC 7.8
Ozone Calibration Routines..................................................................
SEC 7.9
Temperature Calibration Routines........................................................
SEC 7.10 Turbidity Calibration Routines...............................................................
SEC 7.11 Flow Calibration Routines...................................................................
54
55
57
59
60
63
64
76
78
79
82
85
86
88
90
93
95
97
98
100
MODEL 1056
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
SECTION 1.0.
DESCRIPTION AND SPECIFICATIONS
• MULTIPARAMETER INSTRUMENT – single or dual input. Choose from pH/ORP/ISE,
Resistivity/Conductivity, % Concentration, Chlorine, Oxygen, Ozone, Temperature, Turbidity, Flow,
and 420mA Current Input.
• LARGE DISPLAY – large easytoread process measurements.
• EASY TO INSTALL – modular boards, removable connectors, easy to wire power, sensors, and outputs.
• INTUITIVE MENU SCREENS with advanced diagnostics and help screens.
• SEVEN LANGUAGES included: English, French, German, Italian, Spanish, Portuguese, and Chinese.
• HART® AND PROFIBUS® DP Digital Communications options
FEATURES AND APPLICATIONS
The 1056 dualinput analyzer offers single or dual sen
sor input with an unrestricted choice of dual measure
ments. This multiparameter instrument offers a wide
range of measurement choices supporting most indus
trial, commercial, and municipal applications. The
modular design allows signal input boards to be field
replaced making configuration changes easy.
Conveniently, live process values are always displayed
during programming and calibration routines.
QUICK START PROGRAMMING: Exclusive Quick
Start screens appear the first time the 1056 is pow
ered. The instrument autorecognizes each measure
ment board and prompts the user to configure each
sensor loop in a few quick steps for immediate deploy
ment.
DIGITAL COMMUNICATIONS: HART and Profibus
DP digital communications are available. The 1056
HART units communicate with the Model 375 HART®
handheld communicator and HART hosts, such as
AMS Intelligent Device Manager. Model 1056 Profibus
units are fully compatible with Profibus DP networks
and Class 1 or Class 2 masters. HART and Profibus
DP configured units will support any single or dual
measurement configuration of Model 1056.
MENUS: Menu screens for calibrating and programming
are simple and intuitive. Plain language prompts and
help screens guide the user through these procedures.
DUAL SENSOR INPUT AND OUTPUT: The 1056
accepts single or dual sensor input. Standard 0/4
20 mA current outputs can be programmed to
correspond to any measurement or temperature.
ENCLOSURE: The instrument fits standard ½ DIN
panel cutouts. The versatile enclosure design supports
panelmount, pipemount, and surface/wallmount
installations.
ISOLATED INPUTS: Inputs are isolated from other
signal sources and earth ground. This ensures clean
signal inputs for single and dual input configurations.
For dual input configurations, isolation allows any
combination of measurements and signal inputs with
out crosstalk or signal interference.
TEMPERATURE: Most measurements require tem
perature compensation. The 1056 will automatically
recognize Pt100, Pt1000 or 22k NTC RTDs built into
the sensor.
SECURITY ACCESS CODES: Two levels of security
access are available. Program one access code for
routine calibration and hold of current outputs; program
another access code for all menus and functions.
1
MODEL 1056
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
DIAGNOSTICS: The analyzer continuously monitors
itself and the sensor(s) for problematic conditions.
The display flashes Fault and/or Warning when these
conditions occur.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
Diagnostics
Faults
Warnings
Sensor 1
Sensor 2
Out 1: 12.05 mA
Out 2: 12.05 mA
1056012032HT
Instr SW VER: 2.12
AC Freq. Used: 60Hz
Information about
each condition
is quickly accessible
by pressing DIAG on
the keypad. User
help screens are
displayed for most
fault and warning
conditions to assist in
troubleshooting.
DISPLAY: The highcontrast LCD provides live
measurement readouts in large digits and shows up to
four additional process variables or diagnostic
parameters. The display is backlit and the format can
be customized to meet user requirements.
LOCAL LANGUAGES :
Rosemount Analytical extends its worldwide reach by
offering seven local languages – English, French,
German, Italian, Spanish, Portuguese, and Chinese.
Every unit includes user programming menus; calibration
routines; faults and warnings; and user help screens
in all seven languages. The displayed language can
be easily set and changed using the menus.
CURRENT OUTPUTS: Two 420 mA or 020 mA current
outputs are electrically isolated. Outputs are fully scalable
and can be programmed to linear or logarithmic
modes. Output dampening can be enabled with time
constants from 0 to 999 seconds. Output 1 includes
digital signal 420 mA superimposed HART (option HT
only)
2
SPECIAL MEASUREMENTS: The Model 1056 offers
measuring capabilities for many applications.
l Single or Dual Turbidity: Ideal in municipal appli
cations for measurement of lowNTU filtered drinking
water. Must be used with Clarity II sensor, sensor cable
and debubbler.
Model T1056
Clarity® II
Turbidimeter
System
l 4Electrode Conductivity:
The 1056 is compatible with Rosemount
Analytical 4electrode Model 410VP in the PURSENSE
family of conductivity sensors. This sensor supports
a wide array of applications and is capable of measuring
a large range of conductivity with one geometric
configuration. Wired to the 1056, this sensor can
measure 2μS/cm to 300mS/cm with an accuracy of 4%
of reading throughout the entire range.
l 420mA Current Input: Accepts any analog current
input from an external device for temperature compen
sation of measurements and atmospheric pressure
input for partial pressure correction of oxygen.
l Selective Ions: The analyzer is able to measure
ammonia and fluoride using commercially available
ionselective electrodes. All analyzers with installed pH
boards can be programmed to measure selective ions.
l pH Independent Free Chlorine: With Rosemount
Analytical’s 498Cl01 sensor, the analyzer is able to
measure free chlorine with automatic correction for
process pH without the need for a pH sensor.
l Inferential pH: The analyzer is able to derive and
display inferred pH (pHCalc) using two contacting con
ductivity signal boards and the appropriate contacting
conductivity sensors. This method will calculate the
pH of condensate and boiler water from conductivity
and cation conductivity measurements.
l Differential Conductivity: Dual input conductivity
configurations can measure differential conductivity.
The analyzer can be programmed to display dual
conductivity as ratio, % rejection, or % passage.
MODEL 1056
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
SPECIFICATIONS General
Enclosure: Polycarbonate. Type 4X, IP65.
Note: To ensure a watertight seal, tighten all four front
panel screws to 6 inlbs of torque
Power: Code 01: 115/230 VAC ±15%, 50/60 Hz. 10W.
Code 02: 20 to 30 VDC. 15 W.
Code 03: 85 to 265 VAC, 47.5 to 65.0 Hz, switching.
15 W.
Note: Code 02 and 03 power supplies include 4 pro
grammable relays
Equipment protected by double insulation
RFI/EMI: EN61326
LVD:
EN610101
Alarms relays*: Four alarm relays for process meas
urement(s) or temperature. Any relay can be config
ured as a fault alarm instead of a process alarm. Each
relay can be configured independently and each can
be programmed with interval timer settings.
Relays: Form C, SPDT, epoxy sealed
Dimensions: Overall 155 x 155 x 131mm (6.10 x 6.10
x 5.15 in.). Cutout: 1/2 DIN 139mm x 139mm (5.45 x
5.45 in.)
Conduit Openings: Accepts 1/2” or PG13.5 conduit
fittings
Display: Monochromatic graphic liquid crystal display.
128 x 96 pixel display resolution. Backlit. Active
display area: 58 x 78mm (2.3 x 3.0 in.).
Ambient Temperature and Humidity: 0 to 55°C
(32 to 131°F). Turbidity only: 0 to 50°C (32 to
122°F), RH 5 to 95% (noncondensing)
Hazardous Location Approvals
Options for CSA: 01, 02, 03, 20, 21, 22, 24, 25, 26,
27, 30, 31, 32, 34, 35, 36, 37, 38, AN, and HT.
Class I, Division 2, Groups A, B, C, & D
Class Il, Division 2, Groups E, F, & G
Class Ill T4A Tamb= 50° C
Evaluated to the ANSI/UL Standards. The ‘C’ and ‘US’ indi
cators adjacent to the CSA Mark signify that the product has
been evaluated to the applicable CSA and ANSI/UL
Standards, for use in Canada and the U.S. respectively
Options for FM: 01, 02, 03, 20, 21, 22, 24, 25, 26, 30,
31, 32, 34, 35, 36, 38, AN, and HT.
Class I, Division 2, Groups A, B, C, & D
Class Il & lll, Division 2, Groups E, F, & G
T4A Tamb= 50° C Enclosure Type 4X
Storage Temperature Effect: 20 to 60ºC (4 to 140°F)
POLLUTION DEGREE 2: Normally only nonconductive
pollution occurs. Occasionally, however, a temporary
conductivity caused by condensation must be expected.
Altitude: for use up to 2000 meter (6562 ft.)
Maximum Relay Current
Resistive
28 VDC
115 VAC
230 VAC
5.0 A
5.0 A
5.0 A
Inductive load: 1/8 HP motor (max.), 40 VAC
CAUTION
RISK OF ELECTRICAL SHOCK
*Relays only available with 02 power supply (20 30 VDC) or 03
switching power supply (85 265 VAC)
WARNING
WARNING
Exposure to some chemicals may degrade the
sealing properties used in the following devices:
Zettler Relays (K1K4) PN AZ81CH12DSEA
Inputs: One or two isolated sensor inputs
Outputs: Two 420 mA or 020 mA isolated current out
puts. Fully scalable. Max Load: 550 Ohm. Output 1
has superimposed HART signal (configurations
10560X2X3XHT only)
Current Output Accuracy: ±0.05 mA @ 25 ºC
Terminal Connections Rating: Power connector
(3leads): 2412 AWG wire size. Signal board ter
minal blocks: 2616 AWG wire size. Current output
connectors (2leads): 2416 AWG wire size. Alarm
relay terminal blocks: 2412 AWG wire size
(02 24 VDC power supply and 03 85265VAC
power supply)
Weight/Shipping Weight: (rounded up to nearest lb or
nearest 0.5 kg): 3 lbs/4 lbs (1.5 kg/2.0 kg)
3
MODEL 1056
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
CONTACTING CONDUCTIVITY (Codes 20 and 30)
Measures conductivity in the range 0 to 600,000 μS/cm
(600mS/cm). Measurement choices are conductivity,
resistivity, total dissolved solids, salinity, and % concen
tration. The % concentration selection includes the
choice of five common solutions (012% NaOH, 015%
HCl, 020% NaCl, and 025% or 9699.7% H2SO4).
The conductivity concentration algorithms for these
solutions are fully temperature compensated. Three
temperature compensation options are available:
manual slope (X%/°C), high purity water (dilute sodium
chloride), and cation conductivity (dilute hydrochloric
acid). Temperature compensation can be disabled,
allowing the analyzer to display raw conductivity. For
more information concerning the use and operation of
the contacting conductivity sensors, refer to the product
data sheets.
Temperature Specifications:
Temperature range
0150ºC
Temperature Accuracy,
Pt1000, 050 ºC
± 0.1ºC
Temperature Accuracy,
Pt1000, Temp. > 50 ºC
± 0.5ºC
RECOMMENDED SENSORS FOR CONDUCTIVITY:
All Rosemount Analytical ENDURANCE Model 400
series conductivity sensors (Pt 1000 RTD) and
Model 410 sensor.
Note: When two contacting conductivity sensors are
used, Model 1056 can derive an inferred pH value
called pHCalc. pHCalc is calculated pH, not directly
measured pH. (Model 10560X2030AN required)
Note: Selected 4electrode, highrange contacting
conductivity sensors are compatible with Model 1056.
Input filter: time constant 1 999 sec, default 2 sec.
family
4electrode sensors
Response time: 3 seconds to 100% of final reading
Salinity: uses Practical Salinity Scale
Total Dissolved Solids: Calculated by multiplying
conductivity at 25ºC by 0.65
ENDURANCE series of
conductivity sensors
TM
PERFORMANCE SPECIFICATIONS
Recommended Range – Contacting Conductivity
Cell
Constant
0.01
0.1
1.0
4electrode
0.01μS/cm
0.1μS/cm
1.0μS/cm
10μS/cm
100μS/cm
0.01μS/cm to 200μS/cm
1000μS/cm
10mS/cm
100mS/cm
200μS/cm to 6000μS/cm
2000μS/cm to 60mS/cm
0.1μS/cm to 2000μS/cm
1 μS/cm to 20mS/cm
20mS/cm to 600mS/cm
2 μS/cm to 300mS/cm
Cell Constant Linearity
±0.6% of reading in recommended range
+2 to 10% of reading outside high recommended range
4
1000mS/cm
±5% of reading outside low recommended range
±4% of reading in recommended range
MODEL 1056
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
TOROIDAL CONDUCTIVITY (Codes 21 and 31)
Measures conductivity in the range of 1 (one) μS/cm to
2,000,000 μS/cm (2 S/cm), Measurement choices are
conductivity, resistivity, total dissolved solids, salinity,
and % concentration. The % concentration selection
includes the choice of five common solutions (012%
NaOH, 015% HCl, 020% NaCl, and 025% or
9699.7% H2SO4). The conductivity concentration
algorithms for these solutions are fully temperature
compensated. For other solutions, a simpletouse
menu allows the customer to enter his own data. The
analyzer accepts as many as five data points and fits
either a linear (two points) or a quadratic function (three
or more points) to the data. Two temperature compensation
options are available: manual slope (X%/°C) and neutral
salt (dilute sodium chloride). Temperature compensation
can be disabled, allowing the analyzer to display raw
conductivity. Reference temperature and linear temper
ature slope may also be adjusted for optimum results.
For more information concerning the use and operation
of the toroidal conductivity sensors, refer to the product
data sheets.
Temperature Specifications:
Temperature range
25 to 210ºC (13 to 410ºF)
Temperature Accuracy,
Pt100, 25 to 50 ºC
± 0.5ºC
Temperature Accuracy,
Pt100,. 50 to 210ºC
± 1ºC
RECOMMENDED SENSORS:
All Rosemount Analytical submersion/immersion and
flowthrough toroidal sensors.
Repeatability: ±0.25% ±5 μS/cm after zero cal
Input filter: time constant 1 999 sec, default 2 sec.
Response time: 3 seconds to 100% of final reading
Salinity: uses Practical Salinity Scale
Total Dissolved Solids: Calculated by multiplying
conductivity at 25ºC by 0.65
High performance toroidal conductivity sensors
Models 226 and 225
PERFORMANCE SPECIFICATIONS
Recommended Range Toroidal Conductivity
Model
226
1μS/cm
10μS/cm
100μS/cm
1000μS/cm
10mS/cm
100mS/cm
5μS/cm to 500mS/cm
1000mS/cm
2000mS/cm
500mS/cm to 2000mS/cm
225 & 228
15μS/cm to 1500mS/cm
242
222
(1in & 2in)
1500mS/cm to 2000mS/cm
100μS/cm to 2000mS/cm
500μS/cm to 2000mS/cm
LOOP PERFORMANCE (Following Calibration)
Model 226: ±1% of reading ±5μS/cm in recommended range
Models 225 & 228: ±1% of reading ±10μS/cm in recommended range
Models 222,242: ±4% of reading in recommended range
Model 225, 226 & 228: ±5% of reading outside high recommended range
Model 226: ±5μS/cm outside low recommended range
Models 225 & 228: ±15μS/cm outside low recommended range
5
MODEL 1056
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
pH/ORP/ISE (Codes 22 and 32)
For use with any standard pH or ORP sensor.
Measurement choices are pH, ORP, Redox, ammonia,
fluoride or custom ISE. The automatic buffer recognition
feature uses stored buffer values and their temperature
curves for the most common buffer standards available
worldwide. The analyzer will recognize the value of the
buffer being measured and perform a self stabilization
check on the sensor before completing the calibration.
Manual or automatic temperature compensation is
menu selectable. Change in pH due to process temper
ature can be compensated using a programmable tem
perature coefficient. For more information concerning
the use and operation of the pH or ORP sensors, refer
to the product data sheets.
PERFORMANCE SPECIFICATIONS
ANALYZER (ORP INPUT)
Model 1056 can also derive an inferred pH value called
pHCalc (calculated pH). pHCalc can be derived and
displayed when two contacting conductivity sensors are
used. (Model 10560X2030AN)
All standard pH sensors.
Measurement Range [ORP]: 1500 to +1500 mV
Accuracy: ± 1 mV
Temperature coefficient: ±0.12mV / ºC
Input filter: time constant 1 999 seconds, default 4
seconds.
Response time: 5 seconds to 100% of final reading
RECOMMENDED SENSORS FOR pH:
RECOMMENDED SENSORS FOR ORP:
All standard ORP sensors.
PERFORMANCE SPECIFICATIONS
ANALYZER (pH INPUT)
Measurement Range [pH]: 0 to 14 pH
Accuracy: ±0.01 pH
Diagnostics: glass impedance, reference impedance
Temperature coefficient: ±0.002pH/ ºC
Solution temperature correction: pure water, dilute
base and custom.
Buffer recognition: NIST, DIN 19266, JIS 8802, BSI,
DIN19267, Ingold, and Merck.
Input filter: time constant 1 999 seconds, default 4
seconds.
Response time: 5 seconds to 100%
General purpose and high performance pH sensors
Models 396PVP, 399VP and 3300HT
Temperature Specifications:
Temperature range
0150ºC
Temperature Accuracy, Pt100, 050 ºC
± 0.5ºC
Temperature Accuracy, Temp. > 50 ºC
± 1ºC
6
MODEL 1056
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
FLOW (Code 23 and 33)
For use with most pulse signal flow sensors, the 1056
userselectable units of measurement include flow rates
in GPM (Gallons per minute), GPH (Gallon per hour), cu
ft/min (cubic feet per min), cu ft/hour (cubic feet per
hour), LPM (liters per minute), LPH (liters per hour), or
m3/hr (cubic meters per hour), and velocity in ft/sec or
m/sec. When configured to measure flow, the unit also
acts as a totalizer in the chosen unit (gallons, liters, or
cubic meters).
Dual flow instruments can be configured as a % recovery,
flow difference, flow ratio, or total (combined) flow.
Totalized Flow: 0 – 9,999,999,999,999 Gallons or m3,
0 – 999, 999,999,999 cu ft.
Accuracy: 0.5%
Input filter: time constant 0999 sec., default 5 sec.
RECOMMENDED SENSORS*
+GF+ Signet 515 RotorX Flow sensor
* Input voltage not to exceed ±36V
PERFORMANCE SPECIFICATIONS
Frequency Range: 3 to 1000 Hz
Flow Rate: 0 99,999 GPM, LPM, m3/hr, GPH, LPH,
cu ft/min, cu ft/hr.
420mA Current Input (Codes 23 and 33)
For use with any transmitter or external device that
transmits 420mA or 020mA current outputs. Typical
uses are for temperature compensation of live meas
urements (except ORP, turbidity and flow) and for
continuous atmospheric pressure input for determina
tion of partial pressure, needed for compensation of live
dissolved oxygen measurements. External input of
atmospheric pressure for DO measurement allows
continuous partial pressure compensation while the
Model 1056 enclosure is completely sealed. (The
pressure transducer component on the DO board can
only be used for calibration when the case is open to
atmosphere.)
Externally sourced current input is also useful for
calibration of new or existing sensors that require
temperature measurement or atmospheric pressure
inputs (DO only).
For externally sourced temp or pressure compensation,
the user must program the 1056 to input the 420mA
current signal from the external device.
In addition to live continuous compensation of live
measurements, the current input board can also be
used simply to display the measured temperature. or
the calculated partial pressure from the external device.
This feature leverages the large display variables on the
Model 1056 as a convenience for technicians.
Temperature can be displayed in degrees C or degrees
F. Partial pressure can be displayed in inches Hg, mm
Hg, atm (atmospheres), kPa (kiloPascals), bar or mbar.
The current input board can be used with devices that
do not actively power their 420mA output signals. The
Model 1056 actively powers to the + and – lines of the
current input board to enable current input from a
420mA output device.
Note: this Model 1056 signal input board (23, 33
model option code) also includes flow measurement
functionality. The signal board, however, must be
configured to measure either mA current input or flow.
PERFORMANCE SPECIFICATIONS
Measurement Range *[mA]: 020 or 420
Accuracy: ±0.03mA
Input filter: time constant 0999 sec., default 5 sec.
*Current input not to exceed 22mA
7
MODEL 1056
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
CHLORINE (Code 24 and 34)
Free and Total Chlorine
RECOMMENDED SENSORS
The 1056 is compatible with the 499ACL01 free chlorine
sensor and the 499ACL02 total chlorine sensor. The
499ACL02 sensor must be used with the TCL total
chlorine sample conditioning system. The 1056 fully
compensates free and total chlorine readings for
changes in membrane permeability caused by temper
ature changes. For free chlorine measurements, both
automatic and manual pH correction are available. For
automatic pH correction select code 32 and an appropri
ate pH sensor. For more information concerning the use
and operation of the amperometric chlorine sensors and
the TCL measurement system, refer to the product data
sheets.
Rosemount Analytical Model 499ACL03 Monochloramine
sensor
PERFORMANCE SPECIFICATIONS
Resolution: 0.001 ppm or 0.01 ppm – selectable
Input Range: 0nA – 100μA
Automatic pH correction (requires Code 32): 6.0 to
10.0 pH
Temperature compensation: Automatic (via RTD) or
manual (050°C).
Input filter: time constant 1 999 sec, default 5 sec.
Response time: 6 seconds to 100% of final reading
RECOMMENDED SENSORS*
Chlorine: Model 499ACL01 Free Chlorine or Model
499ACL02 Total Residual Chlorine
pH: The following pH sensors are recommended for
automatic pH correction of free chlorine readings:
Models: 3990962, 39914, and 399VP09
pHIndependent Free Chlorine
The 1056 is compatible with the Model 498CL01 pH
independent free chlorine sensor. The Model 498CL01
sensor is intended for the continuous determination of
free chlorine (hypochlorous acid plus hypochlorite ion)
in water. The primary application is measuring chlorine
in drinking water. The sensor requires no acid pretreat
ment, nor is an auxiliary pH sensor required for pH
correction. The Model 1056 fully compensates free
chlorine readings for changes in membrane
permeability caused by temperature. For more information
concerning the use and operation of the amperometric
chlorine sensors, refer to the product data sheets.
PERFORMANCE SPECIFICATIONS
Resolution: 0.001 ppm or 0.01 ppm – selectable
Input Range: 0nA – 100μA
Automatic pH correction: 6.5 to 10.0 pH
Temperature compensation: Automatic (via RTD) or
manual (050°C).
Input filter: time constant 1 999 sec, default 5 sec.
Response time: 6 seconds to 100% of final reading
RECOMMENDED SENSORS
Rosemount Analytical Model 498CL01 pH independent
free chlorine sensor
Monochloramine
The Model 1056 is compatible with the Model 499A CL03
Monochloramine sensor. The Model 1056 fully
compensates readings for changes in membrane
permeability caused by temperature changes. Because
monochloramine measurement is not affected by pH of
the process, no pH sensor or correction is required. For
more information concerning the use and operation of the
amperometric chlorine sensors, refer to the product data
sheets.
PERFORMANCE SPECIFICATIONS
Resolution: 0.001 ppm or 0.01 ppm – selectable
Input Range: 0nA – 100μA
Temperature compensation: Automatic (via RTD) or
manual (050°C).
Input filter: time constant 1 999 sec, default 5 sec.
Response time: 6 seconds to 100% of final reading
8
Chlorine sensors with Variopol connection
and cable connection
Model 498CL01
MODEL 1056
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
DISSOLVED OXYGEN
(Codes 25 and 35)
DISSOLVED OZONE
(Code 26 and 36)
The 1056 is compatible with the 499ADO, 499ATrDO,
Hx438, and Gx438 dissolved oxygen sensors and the
4000 percent oxygen gas sensor. The 1056 displays
dissolved oxygen in ppm, mg/L, ppb, μg/L, % satura
tion, % O2 in gas, ppm O2 in gas. The analyzer fully
compensates oxygen readings for changes in mem
brane permeability caused by temperature changes.
An atmospheric pressure sensor is included on all dis
solved oxygen signal boards to allow automatic atmos
pheric pressure determination at the time of calibration.
If removing the sensor from the process liquid is imprac
tical, the analyzer can be calibrated against a standard
instrument. Calibration can be corrected for process
salinity. For more information on the use of ampero
metric oxygen sensors, refer to the product data
sheets.
The 1056 is compatible with the Model 499AOZ sen
sor. The 1056 fully compensates ozone readings for
changes in membrane permeability caused by temper
ature changes. For more information concerning the
use and operation of the amperometric ozone sensors,
refer to the product data sheets.
PERFORMANCE SPECIFICATIONS
RECOMMENDED SENSOR
Resolution: 0.01 ppm; 0.1 ppb for 499A TrDO sensor
(when O2 <1.00 ppm); 0.1%
Input Range: 0nA – 100μA
Temperature Compensation: Automatic (via RTD) or
manual (050°C).
Input filter: time constant 1 999 sec, default 5 sec.
Response time: 6 seconds to 100% of final reading
Rosemount Analytical Model 499A OZ ozone sensor
PERFORMANCE SPECIFICATIONS
Resolution: 0.001 ppm or 0.01 ppm – selectable
Input Range: 0nA – 100μA
Temperature Compensation: Automatic (via RTD) or
manual (035°C)
Input filter: time constant 1 999 sec, default 5 sec.
Response time: 6 seconds to 100% of final reading
RECOMMENDED SENSORS
Rosemount Analytical amperometric membrane and
steamsterilizable sensors listed above
Dissolved Ozone sensors with Polysulfone body
Variopol connection and cable connection
Model 499AOZ
Dissolved Oxygen sensor with Variopol connection
Model 499ADO
9
MODEL 1056
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
Turbidity (Codes 27 and 37)
The 1056 instrument is available in single and dual tur
bidity configurations for the Clarity II® turbidimeter. It is
intended for the determination of turbidity in filtered
drinking water. The other components of the Clarity II
turbidimeter – sensor(s), debubbler/measuring cham
ber(s), and cable for each sensor must be ordered
separately or as a complete system with the Model
1056.
The 1056 turbidity instrument accepts inputs from both
USEPA 180.1 and ISO 7027compliant sensors
When ordering the Model 1056 turbidity instrument, the
02 (24VDC power supply) or the 03 (switching
115/230VAC power supply) are required. Both of these
power supplies include four fully programmable relays
with timers.
Note: Model 1056 Turbidity must be used with Clarity
II sensor, sensor cable and debubbler.
PERFORMANCE SPECIFICATIONS
Units: Turbidity (NTU, FTU, or FNU); total suspended
solids (mg/L, ppm, or no units)
Display resolutionturbidity: 4 digits; decimal point
moves from x.xxx to xxx.x
Display resolutionTSS: 4 digits; decimal point moves
from x.xxx to xxxx
Calibration methods: userprepared standard, com
mercially prepared standard, or grab sample. For total
suspended solids user must provide a linear calibration
equation.
Inputs: Choice of single or dual input, EPA 180.1 or
ISO 7027 sensors.
Field wiring terminals: removable terminal blocks for
sensor connection.
Accuracy after calibration at 20.0 NTU:
01 NTU ±2% of reading or 0.015 NTU, whichever is
greater.
020 NTU: ±2% of reading.
Clarity ll Turbidimeter
10
MODEL 1056
SECTION 2.0
INSTALLATION
SECTION 2.0.
INSTALLATION
2.1 UNPACKING AND INSPECTION
2.2 INSTALLATION
2.1 UNPACKING AND INSPECTION
Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions. Save the box. If
there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present. If
items are missing, notify Rosemount Analytical immediately.
2.2 INSTALLATION
2.2.1 General Information
1. Although the analyzer is suitable for outdoor use, do not install it in direct sunlight or in areas of extreme tem
peratures.
2. Install the analyzer in an area where vibration and electromagnetic and radio frequency interference are min
imized or absent.
3. Keep the analyzer and sensor wiring at least one foot from high voltage conductors. Be sure there is easy
access to the analyzer.
4. The analyzer is suitable for panel, pipe, or surface mounting. Refer to the table below.
Type of Mounting
Figure
Panel
21
Wall and Pipe
22
WARNING
RISK OF ELECTRICAL SHOCK
Electrical installation must be in accordance with
the National Electrical Code (ANSI/NFPA70)
and/or any other applicable national or local codes.
11
FIGURE 21 PANEL MOUNTING DIMENSIONS
MILLIMETER
INCH
17.13
1.1
126.4
5.0
101.6
4.00
154.9
6.1
154.9
6.1
Front View
Side View
( 126.4
5.0 )
76.2
3.0
41.4
1.6
Bottom View
152.73
6.0
Note: Panel mounting seal integrity (4/4X) for outdoor applications is the responsibility of the end user.
12
FIGURE 22 PIPE AND WALL MOUNTING DIMENSIONS
(Mounting bracket PN:2382000)
MILLIMETER
INCH
154.9
6.1
Wall / Surface Mount
232
9.1
102
4.0
33.5
1.3
130
5.1
187
7.4
154.9
6.1
165
6.5
Side View
Front View
Pipe Mount
232
9.1
Bottom View
33.5
1.3
130
5.1
80.01
3.2
165
6.5
45.21
1.8
108.9
4.3
Side View
71.37
2.8
The front panel is hinged at the bottom. The panel swings down for easy access to the wiring locations.
13
FIGURE 23 CSA NonIncendive Class I, Division 2 Certified product for selected configurations (for approved models, see Fig. 24)
MODEL 1056
14
SECTION 2.0
INSTALLATION
FIGURE 24 CSA NonIncendive Class I, Division 2 Certified product for selected configurations
MODEL 1056
SECTION 2.0
INSTALLATION
15
FIGURE 25 FM NonIncendive Class I, Division 2 Certified product for selected configurations (for approved models, see Fig. 26)
MODEL 1056
16
SECTION 2.0
INSTALLATION
FIGURE 26 FM NonIncendive Class I, Division 2 Certified product for selected configurations
MODEL 1056
SECTION 2.0
INSTALLATION
17
MODEL 1056
18
SECTION 2.0
INSTALLATION
This page left blank intentionally
MODEL 1056
SECTION 3.0
WIRING
SECTION 3.0.
WIRING
3.1
3.2
3.3
3.4
GENERAL
PREPARING CONDUIT OPENINGS
PREPARING SENSOR CABLE
POWER, OUTPUT, AND SENSOR
CONNECTIONS
3.1 GENERAL
The 1056 is easy to wire. It includes removable connectors and slideout signal input boards. The front panel is
hinged at the bottom. The panel swings down for easy access to the wiring locations.
3.1.1. Removable connectors and signal input boards
Model 1056 uses removable signal input boards and communication boards for ease of wiring and instal
lation. Each of the signal input boards can be partially or completely removed from the enclosure for wiring.
The Model 1056 has three slots for placement of up to two signal input boards and one communication
board.
Slot 1Left
Slot 2 – Center
Slot 3 – Right
Comm. board
Input Board 1
Input Board 2
3.1.2. Signal Input boards
Slots 2 and 3 are for signal input measurement boards. Wire the sensor leads to the measurement board
following the lead locations marked on the board. After wiring the sensor leads to the signal board, carefully slide
the wired board fully into the enclosure slot and take up the excess sensor cable through the cable gland. Tighten
the cable gland nut to secure the cable and ensure a sealed enclosure.
3.1.3. Digital Communication boards
HART and Profibus DP communication boards will be available in the future as options for Model 1056 digital
communication with a host. The HART board supports Bell 202 digital communications over an analog
420mA current output. Profibus DP is an open communications protocol which operates over a dedicated
digital line to the host.
3.1.4 Alarm relays
Four alarm relays are supplied with the switching power supply (85 to 265VAC, 03 order code) and the 24VDC
power supply (2030VDC, 02 order code). All relays can be used for process measurement(s) or temperature.
Any relay can be configured as a fault alarm instead of a process alarm. Each relay can be configured
independently and each can be programmed as an interval timer, typically used to activate pumps or control
valves. As process alarms, alarm logic (high or low activation or USP*) and deadband are userprogrammable.
Customerdefined failsafe operation is supported as a programmable menu function to allow all relays to be
energized or notenergized as a default condition upon powering the analyzer.
The USP* alarm can be programmed to activate when the conductivity is within a userselectable
percentage of the limit. USP alarming is available only when a contacting conductivity measurement board is
installed.
3.2 PREPARING CONDUIT OPENINGS
There are six conduit openings in all configurations of Model 1056. (Note that four of the openings will be fitted
with plugs upon shipment.)
Conduit openings accept 1/2inch conduit fittings or PG13.5 cable glands. To keep the case watertight, block
unused openings with NEMA 4X or IP65 conduit plugs.
NOTE: Use watertight fittings and hubs that comply with your requirements. Connect the conduit hub to the
conduit before attaching the fitting to the analyzer.
19
MODEL 1056
SECTION 3.0
WIRING
3.3 PREPARING SENSOR CABLE
The 1056 is intended for use with all Rosemount Analytical sensors. Refer to the sensor installation instructions for
details on preparing sensor cables.
3.4 POWER, OUTPUT, AND SENSOR CONNECTIONS
3.4.1 Power wiring
Three Power Supplies are offered for Model 1056:
a. 115/230VAC Power Supply (01 ordering code)
b. 24VDC (20 – 30V) Power Supply (02 ordering code)
c. 85 – 265 VAC Switching Power Supply (03 ordering code)
AC mains (115 or 230V) leads and 24VDC leads are wired to the Power Supply board which is mounted vertically
on the left side of the main enclosure cavity. Each lead location is clearly marked on the Power Supply board.
Wire the power leads to the Power Supply board using the lead markings on the board.
The grounding plate is connected to the earth terminal of power supply input connector TB1 on the 01
(115/230VAC) and 03 (85265VAC) power supplies. The green colored screws on the grounding plate are intend
ed for connection to some sensors to minimize radio frequency interference. The green screws are not intended
to be used for safety purposes.
115/230VAC Power Supply (01
ordering code) is shown below:
CAUTION
AC Power switch shipped in the 230VAC
position.
Adjust switch upwards to 115VAC position
for 110VAC – 120VAC operation.
Figure 31
24VDC Power Supply (02 ordering code)
is shown below:
This power supply automatically detects DC power and
accepts 20VDC to 30VDC inputs.
Four programmable alarm relays are included.
Figure 32
Switching AC Power Supply (03 ordering
code) is shown below:
This power supply automatically detects AC line condi
tions and switches to the proper line voltage and line
frequency.
Four programmable alarm relays are included.
20
Figure 33
MODEL 1056
SECTION 3.0
WIRING
3.4.2 Current Output wiring
All instruments are shipped with two 420mA current
outputs. Wiring locations for the outputs are on the
Main board which is mounted on the hinged door of the
instrument. Wire the out put leads to the correct posi
tion on the Main board using the lead markings (+/positive,
/negative) on the board. Male mating connectors are
provided with each unit.
Figure 3.4
3.4.3
Alarm relay wiring
Four alarm relays are supplied with the switching power supply (85 to 265VAC, 03 order code) and the 24VDC
power supply (2030VDC, 02 order code). Wire the relay leads on each of the independent relays to the correct
position on the power supply board using the printed lead markings (NO/Normally Open, NC/Normally Closed, or
Com/Common) on the board. See Fig 34.
NO1
COM1
RELAY 1
NC1
NO2
COM2
RELAY 2
NC2
NO3
COM3
RELAY 3
NC3
NO4
COM4
RELAY 4
NC4
Figure 35 Alarm Relay Wiring for Model 1056 Switching Power Supply (03 Order Code)
3.4.4 Sensor wiring to signal boards
Wire the correct sensor leads to the measurement board using the lead locations marked directly on the b o a r d .
After wiring the sensor leads to the signal board, carefully slide the wired board fully into the enclosure slot and
take up the excess sensor cable through the cable gland.
For best EMI/RFI protection use shielded output signal cable enclosed in an earthgrounded metal conduit.
Connect the shield to earth ground. AC wiring should be 14 gauge or greater. Provide a switch or breaker to dis
connect the analyzer from the main power supply. Install the switch or breaker near the analyzer and label it as
the disconnecting device for the analyzer.
Keep sensor and output signal wiring separate from power wiring. Do not run sensor and power wiring in the same
conduit or close together in a cable tray.
WARNING
RISK OF ELECTRICAL SHOCK
Electrical installation must be in accordance with
the National Electrical Code (ANSI/NFPA70)
and/or any other applicable national or local codes.
21
MODEL 1056
SECTION 3.0
WIRING
Sec. 3.4 Signal board wiring
Figure 36 Contacting Conductivity signal board and Sensor cable leads
Figure 37 Toroidal Conductivity Signal board and Sensor cable leads
22
MODEL 1056
SECTION 3.0
WIRING
Figure 38 pH/ORP/ISE signal board and Sensor cable leads
Figure 39 Amperometric signal (Chlorine, Oxygen, Ozone) board and Sensor cable leads
23
MODEL 1056
SECTION 3.0
WIRING
Figure 310 Turbidity signal board with plugin Sensor connection
Figure 311 Flow/Current Input signal board and Sensor cable leads
24
MODEL 1056
SECTION 3.0
WIRING
FIGURE 312 Power Wiring for the 1056 115/230VAC Power Supply (01 Order Code)
FIGURE 313 Power Wiring for the 1056 85265 VAC Power Supply (03 ordering code)
25
MODEL 1056
SECTION 3.0
WIRING
FIGURE 314 Output Wiring for Model 1056 Main PCB
To Main PCB
FIGURE 315 Power Wiring for Model 1056 24VDC Power Supply (02 ordering code)
26
MODEL 1056
SECTION 4.0
DISPLAY AND OPERATION
SECTION 4.0
DISPLAY AND OPERATION
4.1
4.2
4.3
4.4
USER INTERFACE
KEYPAD
MAIN DISPLAY
MENU SYSTEM
4.1 USER INTERFACE
The 1056 has a large display which shows two live
measurement readouts in large digits and up to four
additional process variables or diagnostic parameters
concurrently. The display is backlit and the format can
be customized to meet user requirements. The intu
itive menu system allows access to Calibration, Hold (of
current outputs), Programming, and Display functions by
pressing the MENU button. In addition, a dedicated
DIAGNOSTIC button is available to provide access to
useful operational information on installed sensor(s)
and any problematic conditions that might occur. The
display flashes Fault and/or Warning when these condi
tions occur. Help screens are displayed for most fault
and warning conditions to guide the user in trou
bleshooting.
During calibration and programming, key presses cause
different displays to appear. The displays are self
explanatory and guide the user stepbystep through
the procedure.
4.2 INSTRUMENT KEYPAD
There are 4 Function keys and 4 Selection keys on the
instrument keypad.
Function keys:
The MENU key is used to access menus for program
ming and calibrating the instrument. Four toplevel
menu items appear when pressing the MENU key:
Calibrate: calibrate attached sensors and
analog outputs.
Hold: Suspend current outputs.
Program: Program outputs, measurement,
temperature, security and reset.
Display: Program display format, language,
warnings, and contrast
Pressing MENU always causes the main menu screen
to appear. Pressing MENU followed by EXIT causes
the main display to appear.
27
MODEL 1056
Pressing the DIAG key displays active Faults and
Warnings, and provides detailed instrument information
and sensor diagnostics including: Faults, Warnings,
Sensor 1 and 2 information, Out 1 and Out 2 live current
values, model configuration string e.g. 1056012031
AN, Instrument Software version, and AC frequency
used. Pressing ENTER on Sensor 1 or Sensor 2 pro
vides useful diagnostics and information (as applica
ble): Measurement, Sensor Type, Raw signal value,
Cell constant, Zero Offset, Temperature, Temperature
SECTION 4.0
DISPLAY AND OPERATION
Offset, selected measurement range, Cable
Resistance, Temperature Sensor Resistance, Signal
Board software version.
The ENTER key. Pressing ENTER stores numbers and
settings and moves the display to the next screen.
The EXIT key. Pressing EXIT returns to the previous
screen without storing changes.
Selection keys:
Surrounding the ENTER key, four Selection keys – up,
down, right and left, move the cursor to all areas of the
screen while using the menus.
Selection keys are used to:
1. select items on the menu screens
2. scroll up and down the menu lists.
3. enter or edit numeric values.
4. move the cursor to the right or left
5. select measurement units during operations
4.3 MAIN DISPLAY
The Model 1056 displays one or two primary measurement
values, up to four secondary measurement values, a
fault and warning banner, alarm relay flags, and a
digital communications icon.
Process measurements:
Two process variables are displayed if two signal boards are installed. One process variable and process tem
perature is displayed if one signal board is installed with one sensor. The Upper display area shows the Sensor
1 process reading. The Center display area shows the Sensor 2 process reading. For dual conductivity, the Upper
and Center display areas can be assigned to different process variables as follows:
Process variables for Upper display example:
Process variables for Center display example:
Measure 1
Measure 1
% Reject
Measure 2
% Pass
% Reject
Ratio
% Pass
Ratio
Blank
For single input configurations, the Upper display area
shows the live process variable and the Center display
area can be assigned to Temperature or blank.
Secondary values:
Up to four secondary values are shown in four display
quadrants at the bottom half of the screen. All four
secondary value positions can be programmed by the
user to any display parameter available. Possible
secondary values include:
Displayable Secondary Values
Slope 1
Man Temp 2
Ref Off 1
Output 1 mA
Gl Imp 1
Output 2 mA
Ref Imp 1
Output 1 %
Raw
Output 2 %
mV Input
Measure 1
Temp 1
Blank
Man Temp 1
28
MODEL 1056
SECTION 4.0
DISPLAY AND OPERATION
Fault and Warning banner:
If the analyzer detects a problem with itself or the sensor the word Fault or Warning will appear at the bottom of
the display. A fault requires immediate attention. A warning indicates a problematic condition or an impending fail
ure. For troubleshooting assitance, press Diag.
Formatting the Main Display
The main display screen can be programmed to show primary process variables, secondary process variables and
diagnostics.
1.
Press MENU
2.
Scroll down to Display. Press ENTER.
3.
Main Format will be highlighted. Press ENTER.
4.
The sensor 1 process value will be highlighted in reverse video. Press the selection keys to navigate down
to the screen sections that you wish to program. Press ENTER.
5.
Choose the desired display parameter or diagnostic for each of the four display sections in the lower screen.
6.
Continue to navigate and program all desired screen sections.
return to the main display.
Press MENU and EXIT. The screen will
For single sensor configurations, the default display shows the live process measurement in the upper display area
and temperature in the center display area. The user can elect to disable the display of temperature in the cen
ter display area using the Main Format function. See Fig. 41 to guide you through programming the main display
to select process parameters and diagnostics of your choice.
For dual sensor configurations, the default display shows Sensor 1 live process measurement in the upper display
area and Sensor 2 live process measurement temperature in the center display area. See Fig. 41 to guide you
through programming the main display to select process parameters and diagnostics of your choice.
4.4 MENU SYSTEM
Model 1056 uses a scroll and select menu system.
Pressing the MENU key at any time opens the toplevel
menu including Calibrate, Hold, Program and Display
functions.
To find a menu item, scroll with the up and down keys
until the item is highlighted. Continue to scroll and
select menu items until the desired function is chosen.
To select the item, press ENTER. To return to a previ
ous menu level or to enable the main live display,
press the EXIT key repeatedly. To return immediately
to the main display from any menu level, simply press
MENU then EXIT.
The selection keys have the following functions:
The Up key (above ENTER) increments numerical values, moves the decimal place one place to the right,
or selects units of measurement.
The Down key (below ENTER) decrements numerical values, moves the decimal place one place to the
left, or selects units of measurement
The Left key (left of ENTER) moves the cursor to the left.
The Right key (right of ENTER) moves the cursor to the right.
To access desired menu functions, use the “Quick Reference” Figure B. During all menu displays (except main
display format and Quick Start), the live process measurements and secondary measurement values are
displayed in the top two lines of the Upper display area. This conveniently allows display of the live values during
important calibration and programming operations.
Menu screens will time out after two minutes and return to the main live display.
29
MODEL 1056
SECTION 4.0
DISPLAY AND OPERATION
FIGURE 41 Formatting the Main Display
30
MODEL 1056
SECTION 5.0
PROGRAMMING THE ANALYZER BASICS
SECTION 5.0.
PROGRAMMING THE ANALYZER BASICS
5.1 GENERAL
5.2 CHANGING STARTUP SETTINGS
5.3 PROGRAMMING TEMPERATURE
5.4 CONFIGURING AND RANGING 420MA OUTPUTS
5.5 SETTING SECURITY CODES
5.6 SECURITY ACCESS
5.7 USING HOLD
5.8 RESETTING FACTORY DEFAULTS – RESET ANALYZER
5.9 PROGRAMMING ALARM RELAYS
5.1 GENERAL
Section 5.0 describes the following programming functions:
Changing the measurement type, measurement units and temperature units.
Choose temperature units and manual or automatic temperature compensation mode
Configure and assign values to the current outputs
Set a security code for two levels of security access
Accessing menu functions using a security code
Enabling and disabling Hold mode for current outputs
Choosing the frequency of the AC power (needed for optimum noise rejection)
Resetting all factory defaults, calibration data only, or current output settings only
5.2 CHANGING STARTUP SETTINGS
5.2.1 Purpose
To change the measurement type, measurement units, or temperature units that were initially entered in Quick
Start, choose the Reset analyzer function (Sec. 5.9) or access the Program menus for sensor 1 or sensor 2 (Sec.
6.0). The following choices for specific measurement type, measurement units are available for each sensor meas
urement board.
TABLE 51. Measurements and Measurement Units
Signal board
pH/ORP (22, 32)
Contacting conductivity
(20, 30)
Toroidal conductivity
(21, 31)
Available measurements
Measurements units:
pH, ORP, Redox, Ammonia, Fluoride,
Custom ISE
Conductivity, Resistivity, TDS, Salinity,
NaOH (012%), HCl (015%), Low H2SO4,
High H2SO4, NaCl (020%),
Custom Curve
pH, mV (ORP)
%, ppm, mg/L, ppb, μg/L, (ISE)
Conductivity, Resistivity, TDS, Salinity,
NaOH (012%), HCl (015%), Low H2SO4,
High H2SO4, NaCl (020%),
Custom Curve
Chlorine
(24, 34)
Oxygen
(25, 35)
Free Chlorine, pH Independ. Free Cl, Total
Chlorine, Monochloramine
Ozone (26, 36)
Temperature (all)
Ozone
Temperature
Oxygen (ppm), Trace Oxygen (ppb),
Percent Oxygen in gas, Salinity
μS/cm, mS/cm, S/cm
% (concentration)
μS/cm, mS/cm, S/cm
% (concentration)
ppm, mg/L
ppm, mg/L, ppb, µg/L % Sat, Partial
Pressure, % Oxygen In Gas, ppm
Oxygen In Gas
ppm, mg/L, ppb, μg/L
°C. ºF
5.2.2 Procedure.
Follow the Reset Analyzer procedure (Sec 5.8) to reconfigure the analyzer to display new measurements or
measurement units. To change the specific measurement or measurement units for each signal board type,
refer to the Program menu for the appropriate measurement (Sec. 6.0).
31
MODEL 1056
SECTION 5.0
PROGRAMMING THE ANALYZER BASICS
5.3 CHOOSING TEMPERATURE UNITS AND AUTOMATIC/MANUAL TEMPERATURE
COMPENSATION
5.3.1 Purpose
Most liquid analytical measurements (except ORP)
require temperature compensation. The Model 1056
performs temperature compensation automatically by
applying internal temperature correction algorithms.
Temperature correction can also be turned off. If tem
perature correction is off, the Model 1056 uses the tem
perature entered by the user in all temperature correc
tion calculations.
5.3.2 Procedure.
Follow the menu screens in Fig. 5.1 to select automatic
or manual temp compensation, set the manual
reference temperature, and to program temperature
units as °C or °F.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
Temperature
Units:
°C
S1 Temp Comp: Auto
S2 Temp Comp: Auto
S1 Manual:
+25.0°C
S2 Manual:
+25.0ºC
Figure 51. Choosing Temp Units and Manual Auto Temp Compensation
5.4 CONFIGURING AND RANGING THE CURRENT OUTPUTS
5.4.1 Purpose
The Model 1056 accepts inputs from two sensors and
has two analog current outputs. Ranging the outputs
means assigning values to the low (0 or 4 mA) and high
(20 mA) outputs. This section provides a guide for
configuring and ranging the outputs. ALWAYS
CONFIGURE THE OUTPUTS FIRST.
5.4.2 Definitions
1. CURRENT OUTPUTS. The analyzer provides a con
tinuous output current (420 mA or 020 mA) directly
proportional to the process variable or temperature.
32
The low and high current outputs can be set to any
value.
2. ASSIGNING OUTPUTS. Assign a measurement to
Output 1 or Output 2.
3. DAMPEN. Output dampening smooths out noisy
readings. It also increases the response time of the
output. Output dampening does not affect the
response time of the display.
4. MODE. The current output can be made directly
proportional to the displayed value (linear mode) or
directly proportional to the common logarithm of the
displayed value (log mode).
MODEL 1056
SECTION 5.0
PROGRAMMING THE ANALYZER BASICS
5.4.3 Procedure: Configure Outputs.
Under the Program/Outputs menu, the adjacent screen
will appear to allow configuration of the outputs. Follow
the menu screens in Fig. 52 to configure the outputs.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
OutputM Configure
Assign:
S1 Meas
Range:
420mA
Scale:
Linear
Dampening:
0sec
Fault Mode: Fixed
Fault Value: 21.00mA
5.4.4 Procedure: Assigning Measurements the Low
and High Current Outputs
The adjacent screen will appear when entering the
Assign function under Program/Output/Configure.
These screens allow you to assign a measurement,
process value, or temperature input to each output.
Follow the menu screens in Fig. 52 to assign
measurements to the outputs.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
5.4.5 Procedure: Ranging the Current Outputs
The
adjacent
screen
will
appear
under
Program/Output/Range. Enter a value for 4mA and
20mA (or 0mA and 20mA) for each output. Follow the
menu screens in Fig. 52 to assign values to the out
puts.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
OutputM Assign
S1 Measurement
S1 Temperature
S2 Measurement
S2 Temperature
OM
OM
OM
OM
SN
SN
SN
SN
Output Range
4mA: 0.000µS/cm
20mA: 20.00µS/cm
4mA: 00.00pH
20mA: 14.00pH
Figure 52. Configuring and Ranging the Current Outputs
33
MODEL 1056
SECTION 5.0
PROGRAMMING THE ANALYZER BASICS
2.
3.
5.5 SETTING A SECURITY CODE
5.5.1 Purpose.
The security codes prevent accidental or unwanted
changes to program settings, displays, and calibration.
Model 1056 has two levels of security code to control
access and use of the instrument to different types of
users. The two levels of security are:
All: This is the Supervisory security level. It
allows access to all menu functions, including
Programming, Calibration, Hold and Display.
Calibration/Hold: This is the operator or
technician level menu. It allows access to
only calibration and Hold of the current outputs.
5.5.2 Procedure.
1. Press MENU. The main menu screen appears.
Choose Program.
Scroll down to Security. Select Security.
The security entry screen appears. Enter a
three digit security code for each of the desired
security levels. The security code takes effect
two minutes after the last key stroke.
Record
the security code(s) for future access
and
communication to operators or techni
cians as
needed.
4.
The display returns to the security menu
screen. Press EXIT to return to the previous
screen. To return to the main display, press
MENU followed by EXIT.
Fig. 53 displays the security code screens.
Program
MAIN MENU
Figure 53. Setting a Security Code
S1: 1.234µS/cm
S2: 12.34pH
25.0ºC
25.0ºC
Program
Outputs
Measurement
Temperature
S1: 1.234µS/cm
S2: 12.34pH
Security
Diagnostic Setup
Ambient AC Power:Unk
Reset Analyzer
34
25.0ºC
25.0ºC
Security
Calibration/Hold: 000
All: 000
MODEL 1056
SECTION 5.0
PROGRAMMING THE ANALYZER BASICS
5.6 SECURITY ACCESS
5.7 USING HOLD
5.6.1 How the Security Code Works
When entering the correct access code for the
Calibration/Hold security level, the Calibration and
Hold menus are accessible. This allows operators or
technicians to perform routine maintenance. This
security level does not allow access to the Program or
Display menus.
When entering the correct access code for All security
level, the user has access to all menu functions, includ
ing Programming, Calibration, Hold and Display.
5.7.1 Purpose
The analyzer output is always proportional to measured
value. To prevent improper operation of systems or
pumps that are controlled directly by the current
output, place the analyzer in hold before removing
the sensor for calibration and maintenance. Be sure
to remove the analyzer from hold once calibration is
complete. During hold, both outputs remain at the last
value. Once in hold, all current outputs remain on
Hold indefinitely.
5.6.2 Procedure.
1. If a security code has been programmed, selecting
the Calibrate, Hold, Program or Display top menu
items causes the security access screen to appear
5.7.2 Using the Hold Function
To hold the outputs,
1.
Press MENU. The main menu screen appears.
Choose Hold.
2.
The Hold Outputs and Alarms? screen
appears. Choose Yes to place the analyzer in
hold. Choose No to take the analyzer out of
hold.
Note: There are no alarm relays with this con
figuration. Current outputs are included with all
configurations.
3.
The Hold screen will then appear and Hold
will remain on indefinitely until Hold is
disabled.
2. Enter the threedigit security code for the appropriate
security level.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
Security Code
000
3. If the entry is correct, the appropriate menu screen
appears. If the entry is incorrect, the Invalid Code
screen appears. The Enter Security Code screen
reappears after 2 seconds.
See figure 51 below.
Hold
MAIN MENU
Figure 54. Using Hold
S1: 1.234µS/cm
S2: 12.34pH
25.0ºC
25.0ºC
Hold
S1 Hold:
S2 Hold:
No
No
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
S1 Hold outputs
and alarms?
No
Yes
35
MODEL 1056
SECTION 5.0
PROGRAMMING THE ANALYZER BASICS
5.8 RESETTING FACTORY DEFAULT SETTINGS
5.8.1 Purpose.
This section describes how to restore factory calibration and default values. The process also clears all fault messages
and returns the display to the first Quick Start screen. The Model 1056 offers three options for resetting factory
defaults.
a. reset all settings to factory defaults
b. reset sensor calibration data only
c. reset analog output settings only
5.8.2. Procedure.
To reset to factory defaults, reset calibration data only or reset analog outputs only, follow the Reset Analyzer flow
diagram.
Figure 55. Resetting Factory Default Settings
36
MODEL 1056
SECTION 5.0
PROGRAMMING THE ANALYZER BASICS
5.9 Programming Alarm Relays
5.9.1 Purpose.
The Model 1056 24VDC (02 order code) and the AC switching power supply (03 order code) provide four alarm
relays for process measurement or temperature. Each alarm can be configured as a fault alarm instead of a
process alarm. Also, each relay can be programmed independently and each can be programmed as an interval
timer. This section describes how to configure alarm relays, simulate relay activation, and synchronize timers for
the four alarm relays. This section provides details to program the following alarm features:
Sec.
Alarm relay feature:
default
Description
5.9.2
Enter Setpoint
100.0uS/cm
Enter alarm trigger value
5.9.3
Assign measurement S1 Measure
Select alarm assignment
5.9.4
Set relay logic
Program relay to activate at High or Low reading
High
0.00uS/cm
Program the change in process value after the relay deactivates
5.9.5
Deadband:
5.9.6
USP Safety:
5.9.7
Normal state:
5.9.8
Interval time:
24.0 hr
Time in hours between relay activations
5.9.9
OnTime:
10 min
Enter the time in seconds that the relay is activated.
5.9.10
Recover time:
60 sec
Enter time after the relay deactivation for process recovery
5.9.11
Hold while active:
5.9.12
Simulate
5.9.13
Synchronize Timers
0%↓
Open
S1
Program percentage of the limit to activate the alarm
Program relay default condition as open or closed for failsafe operation
Holds current outputs during relay activation
Manually simulate alarms to confirm relay operation
Yes
Control the timing of two or more relay timers set as Interval timers
Under the Program/Alarms menu, this screen will
appear to allow configuration of the alarm relays.
Follow the menu screens in Fig. XX to configure the
outputs.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
This screen will appear to allow selection of a specific
alarm relay. Select the desired alarm and press
ENTER.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
This screen will appear next to allow complete pro
gramming of each alarm. Factory defaults are dis
played as they would appear for an installed contact
ing conductivity board. USP Safety only appears if
alarm logic is set to “USP”. Interval timer, On Time,
Recover Time, and Hold While Active only appear if
the alarm is configured as an Interval timer.
Alarms
Configure/Setpoint
Simulate
Synchronize Timers: Yes
Configure/Setpoint
Alarm 1
Alarm 2
Alarm 3
Alarm 4
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
AlarmM Settings
Setpoint:
100.0uS/cm
Assign:
S1 Measure
Logic:
High
Deadband: 0.00uS/cm
USP Safety:
0%↓
Interval time:
24.0 hr
On Time:
120 sec
Recover time:
60 sec
Hold while active: Sens1
37
MODEL 1056
5.9.2 Procedure – Enter Setpoints
Under the Program/Alarms menu, this screen will
appear to allow configuration of the alarm relays.
Enter the desired value for the process measurement
or temperature at which to activate an alarm event.
5.9.3 Procedure – Assign Measurement
Under the Alarms Settings menu, this screen will
appear to allow assignment of the alarm relays. select
an alarm assignment. Additional assignment choices
are shown in Figure XX depending on which meas
urement board(s) is installed.
SECTION 5.0
PROGRAMMING THE ANALYZER BASICS
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
Alarm1 S2 Setpoint
+100.0uS/cm
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
AlarmM Assign:
S1 Measurement
S1 Temperature
S2 Measurement
S2 Temperature
Interval Timer
Fault
Off
5.9.4 Procedure – Set Relay Logic
Under the Alarms Settings menu, this screen will
appear to set the alarm logic. Select the desired relay
logic to activate alarms at a High reading or a Low
reading. USP Safety only appears if a contacting con
ductivity board is installed.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
5.9.5 Procedure – Deadband
Under the Alarms Settings menu, this screen will
appear to program the deadband as a measurement
value. Enter the change in the process value needed
after the relay deactivates to return to normal (and
thereby preventing repeated alarm activation).
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
5.9.6 Procedure – USP Safety
Under the Alarms Settings menu, this screen will
appear to program the USP alarm setting. Enter the
percentage below the limit at which to activate the
alarm.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
5.9.7 Procedure – Normal state
The user can define failsafe condition in software by
programming the alarm default state to normally open
or normally closed upon power up. To display this
alarm configuration item, enter the Expert menus by
holding down the EXIT key for 6 seconds while in the
main display mode. Select Yes upon seeing the screen
prompt: “Enable Expert Menu?”
Under the Alarms Settings menu, this screen will
appear to set the normal state of the alarms. Select the
alarm condition that is desired each time the analyzer is
powering up.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
38
AlarmM Logic:
High
Low
USP
Alarm1 Deadband
+000.5uS/cm
Alarm1 USP Safety
+0%i
Alarm2 Normal State
Open
Closed
MODEL 1056
SECTION 5.0
PROGRAMMING THE ANALYZER BASICS
5.9.8 Procedure – Interval time
Under the Alarms Settings menu, this screen will
appear to set the interval time. Enter the fixed time in
hours between relay activations.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
5.9.9 Procedure – On time
Under the Alarms Settings menu, this screen will
appear to set the relay on time. Enter the time in sec
onds that the relay is activated.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
5.9.10 Procedure – Recovery time
Under the Alarms Settings menu, this screen will
appear to set the relay recovery time. Enter time after
the relay deactivation for process recovery.
Alarm1 Interval Time
024.0 hrs
Alarm1 OnTime
00.00sec
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
Alarm1 Recovery
060sec
5.9.11 Procedure – Hold while active
Under the Alarms Settings menu, this screen will
appear to program the feature that Holds the current
outputs while alarms are active. Select to hold the
current outputs for Sensor 1, Sensor 2 or both sensors
while the relay is activated.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
5.9.12 Procedure – Simulate
Alarm relays can be manually set for the purposes of
checking devices such as valves or pumps. Under the
Alarms Settings menu, this screen will appear to allow
manual forced activation of the alarm relays. Select
the desired alarm condition to simulate.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
5.9.13 Procedure – Synchronize
Under the Alarms Settings menu, this screen will
appear to allow Synchronization of alarms that are set
to interval timers. Select yes or no to Synchronize
two or more timers.
Alarm1 Hold while active
Sensor 1
Sensor 2
Both
None
Simulate Alarm M
Don’t simulate
Deenergize
Energize
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
Synchronize Timers
Yes
No
39
MODEL 1056
40
SECTION 5.0
PROGRAMMING THE ANALYZER BASICS
This page left blank intentionally
MODEL 1056
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
SECTION 6.0
PROGRAMMING MEASUREMENTS
6.1
6.2
6.3
6.4
6.5
6.6
CONFIGURING MEASUREMENTS – INTRODUCTION
pH
ORP
CONTACTING CONDUCTIVITY
TOROIDAL CONDUCTIVITY
CHLORINE
6.6.1 FREE CHLORINE
6.6.2 TOTAL CHLORINE
6.6.3 MONOCHLORAMINE
6.6.4 pHINDEPENDENT FREE CHLORINE
6.7 OXYGEN
6.8 OZONE
6.9 TURBIDITY
6.10 FLOW
6.11 CURRENT INPUT
6.1 PROGRAMMING MEASUREMENTS – INTRODUCTION
The Model 1056 automatically recognizes each installed measurement board upon first powerup and each time
the analyzer is powered. Completion of Quick Start screens upon first power up enable measurements, but addi
tional steps may be required to program the analyzer for the desired measurement application. This section cov
ers the following programming and configuration functions;
1. Selecting measurement type or sensor type (all sections)
2. Identifying the preamp location (pHsee Sec. 6.2)
3. Enabling manual temperature correction and entering a reference temperature (all sections)
4. Enabling sample temperature correction and entering temperature correction slope (selected sections)
5. Defining measurement display resolution (pH and amperometric)
6. Defining measurement display units (all sections)
7. Adjusting the input filter to control display and output reading variability or noise (all sections)
8. Selecting a measurement range (conductivity – see Sec’s 6.4, 6.5)
9. Entering a cell constant for a contacting or toroidal sensor (see Sec’s 6.4, 6.5)
10. Entering a temperature element/RTD offset or temperature slope (conductivitysee Sec’s 6.4)
11. Creating an applicationspecific concentration curve (conductivitysee Sec’s 6.4, 6.5)
12. Enabling automatic pH correction for free chlorine measurement (Sec. 6.6.1)
To fully configure the analyzer for each installed measurement board, you may use the following:
1. Reset Analyzer function to reset factory defaults and configure the measurement board to the desired
measurement. Follow the Reset Analyzer menu (Fig. 55) to reconfigure the analyzer to display new
measurements or measurement units.
2. Program menus to adjust any of the programmable configuration items. Use the following configuration
and programming guidelines for the applicable measurement.
41
MODEL 1056
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
6.2 pH MEASUREMENT PROGRAMMING
6.2.1 Description
This section describes how to configure the Model 1056 analyzer for pH measurements. The following programming and
configuration functions are covered.
TABLE 61. pH Measurement Programming
Measure
pH
default setting
Description
Sec.
Menu function:
6.2.2
Measurement type:
6.2.3
Preamp location:
6.2.4
Solution temperature correction Off
Select Off, ultrapure, high pH, custom
6.2.5
Temp coefficient
Enter the temp coefficient
6.2.6
Resolution:
6.2.7
Filter:
6.2.8
Reference Z:
pH
Analyzer
(custom)
0.01pH
4 sec
Low
Select pH, ORP, Redox, Ammonia, Fluoride, Custom ISE
Identify preamp location
Select 0.01pH or 0.1pH for pH display resolution
Override the default input filter, enter 0999 seconds
Select low or high reference impedance
A detailed flow diagram for pH programming is provided at the end of Sec. 6 to guide you through
all basic programming and configuration functions.
To configure the pH measurement board:
1. Press MENU
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to
pH. Press ENTER.
The adjacent screen format will appear (factory defaults
are shown). To program any function, scroll to the
desired item and press ENTER.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Configure
Measure:
pH
Preamp:
Analyzer
Sol’n Temp Corr: Off
T Coeff: 0.029pH/°C
Resolution: 0.01pH
Filter:
4 sec
Reference Z:
Low
The following subsections provide you with the initial display screen that appears for each configuration function.
Use the flow diagram for pH programming at the end of Sec. 6 and the Model 1056 live screen prompts for each
function to complete configuration and programming.
6.2.2 Measurement
The display screen for selecting the measurement is
shown. The default value is displayed in bold type.
Refer to the pH/ORP Programming flow diagram to
complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Measurement
pH
ORP
Redox
Ammonia
Fluoride
Custom ISE
6.2.3 Preamp
The display screen for identifying the Preamp location is
shown. The default value is displayed in bold type.
Refer to the pH/ORP Programming flow diagram to
complete this function.
42
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Preamp
Analyzer
Sensor/JBox
MODEL 1056
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
6.2.4 Solution Temperature Correction
The display screen for selecting the Solution
temperature correction algorithm is shown. The default
value is displayed in bold type. Refer to the pH/ORP
Programming flow diagram to complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Sol’n Temp Corr.
Off
Ultra Pure Water
High pH
Custom
6.2.5 Temperature Coefficient
The display screen for entering the custom solution tem
perature coefficient is shown. The default value is dis
played in bold type.
Refer to the pH/ORP
Programming flow diagram to complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Sol’n Temp Coeff.
0.032pH/ºC
6.2.6 Resolution
The display screen for selecting 0.01pH or 0.1pH for pH
display resolution is shown. The default value is displayed
in bold type. Refer to the pH/ORP Programming flow
diagram to complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Resolution
0.01pH
0.1pH
6.2.7 Filter
The display screen for entering the input filter value in
seconds is shown. The default value is displayed in
bold type. Refer to the pH/ORP Programming flow diagram
to complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Input filter
04 sec
6.2.8 Reference Impedence
The display screen for selecting Low or High Reference
impedance is shown. The default value is displayed in
bold type. Refer to the pH/ORP Programming flow diagram
to complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Reference Z
Low
High
6.3 ORP MEASUREMENT PROGRAMMING
6.3.1 Description
The section describes how to configure the Model 1056 analyzer for ORP measurements. The following programming
and configuration functions are covered:
TABLE 62. ORP Measurement Programming
Measure
ORP
Sec.
Menu function:
6.3.2
Measurement type:
6.3.3
Preamp location:
6.3.4
Filter:
6.3.5
Reference Z:
default
pH
Analyzer
4 sec
Low
Description
Select pH, ORP, Redox, Ammonia, Fluoride, Custom ISE
Identify preamp location
Override the default input filter, enter 0999 seconds
Select low or high reference impedance
43
MODEL 1056
A detailed flow diagram for ORP programming is
provided at the end of Sec. 6 to guide you through
all basic programming and configuration functions.
To configure the ORP measurement board:
1. Press MENU
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to
ORP. Press ENTER.
The adjacent screen format will appear (factory defaults
are shown). To program any displayed function, scroll
to the desired item and press ENTER.
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Configure
Measure:
pH
Preamp: Analyzer
Flter:
4 sec
Reference Z: Low
The following subsections provide you with the initial display screen that appears for each configuration function.
Use the flow diagram for ORP programming at the end of Sec. 6 and the Model 1056 live screen prompts for
each function to complete configuration and programming.
6.3.2 Measurement
The display screen for selecting the measurement is
shown. The default value is displayed in bold type.
Refer to the pH/ORP Programming flow diagram to
complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
6.3.3 Preamp
The display screen for identifying the Preamp location is
shown. The default value is displayed in bold type.
Refer to the pH/ORP Programming flow diagram to
complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
6.3.4 Filter
The display screen for entering the input filter value in
seconds is shown. The default value is displayed in
bold type. Refer to the pH/ORP Programming flow diagram
to complete this function.
6.3.5 Reference Impedence
The display screen for Selecting Low or high Reference
impedance is shown. The default value is displayed in
bold type. Refer to the pH/ORP Programming flow diagram
to complete this function.
44
SN Measurement
pH
ORP
Redox
Ammonia
Fluoride
Custom ISE
SN Preamp
Analyzer
Sensor/JBox
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Input filter
04 sec
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Reference Z
Low
High
MODEL 1056
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
6.4 CONTACTING CONDUCTIVITY MEASUREMENT PROGRAMMING
6.4.1 Description
The section describes how to configure the Model 1056 analyzer for conductivity measurements using contacting
conductivity sensors. The following programming and configuration functions are covered.
TABLE 63. Contacting Conductivity Measurement Programming
Measure
Contacting
Conductivity
default
Description
Sec.
Menu function:
6.4.2
Type:
6.4.3
Measure:
6.4.4
Range:
Auto
6.4.5
Cell K:
1.00000/cm
6.4.6
RTD Offset:
0.00ºC
Enter the RTD Offset
6.4.7
RTD Slope:
0
Enter the RTD Slope
6.4.8
Temp Comp:
6.4.9
Slope:
6.4.10
Ref Temp:
6.4.11
Filter:
6.4.12
Custom
6.4.13
Cal Factor:
2Electrode
Conductivity
Slope
2.00%/°C
25.0°C
2 sec
Setup
0.95000/cm
Select 2Electrode or 4Electrode type sensors
Select Conductivity, Resistivity, TDS. Salinity or % conc
Select measurement Autorange or specific range
Enter the cell Constant for the sensor
Select Temp Comp: Slope, Neutral Salt, Cation or Raw
Enter the linear temperature coefficient
Enter the Reference temp
Override the default input filter, enter 0999 seconds
Enter 25 data points in ppm and µS/cm for custom curves
Enter the Cal Factor for 4Electrode sensors from the sensor tag
A detailed flow diagram for contacting conductivity programming is provided at the end of Sec. 6 to
guide you through all basic programming and configuration functions.
To configure the contacting conductivity measurement
board:
1. Press MENU
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to
contacting conductivity. Press ENTER.
The adjacent screen format will appear (factory defaults
are shown). To program any displayed function, scroll
to the desired item and press ENTER.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Configure
Type:
2Electrode
Measure:
Cond
Range:
Auto
Cell K: 1.00000/cm
RTD Offset: 0.00ºC
RTD Slope:
0
Temp Comp: Slope
Slope:
2.00%/°C
Ref Temp:
25.0°C
Filter:
2 sec
Custom Setup
The following subsections provide you with the initial display screen that appears for each configuration function.
Use the flow diagram for contacting conductivity programming at the end of Sec. 6 and the Model 1056 live
screen prompts for each function to complete configuration and programming.
6.4.2 Sensor Type
The display screen for selecting 2Electrode or
4Electrode type sensors is shown. The default value
is displayed in bold type. Refer to the contacting
conductivity Programming flow diagram to complete this
function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Type
2Electrode
4Electrode
45
MODEL 1056
6.4.3 Measure
The display screen for selecting the measurement is
shown. The default value is displayed in bold type.
Refer to the contacting conductivity Programming flow
diagram to complete this function.
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Measurement
Conductivity
Resistivity
TDS
Salinity
NaOH (012%)
HCl (015%)
Low H2SO4
High H2SO4
NaCl (020%)
Custom Curve
6.4.4 Range
The display screen for Selecting Autoranging or a specific
range is shown. The default value is displayed in bold
type. Note: Ranges are shown as conductance, not
conductivity. Refer to the contacting conductivity
Programming flow diagram to complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
6.4.5 Cell Constant
The display screen for entering a cell Constant for the
sensor is shown. The default value is displayed in bold
type. Refer to the contacting conductivity Programming
flow diagram to complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
6.4.6 RTD Offset
The display screen for Entering the RTD Offset for the
sensor is shown. The default value is displayed in bold
type. Refer to the contacting conductivity Programming
flow diagram to complete this function.
SN Range
Auto
50 µS
500 µS
2000 µS
20 mS
200 mS
600 mS
SN Cell Constant
1.00000 /cm
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN RTD Offset
0.00°C
6.4.7 RTD Slope
The display screen for entering the RTD slope for the
sensor is shown. The default value is displayed in bold
type. Refer to the contacting conductivity Programming
flow diagram to complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
6.4.8 Temp Comp
The display screen for Selecting Temperature
Compensation as Slope, Neutral Salt, Cation or Raw is
shown. The default value is displayed in bold type.
Refer to the contacting conductivity Programming flow
diagram to complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
46
SN RTD Slope
2.00%/ºC
SN Temp Comp
Slope
Neutral Salt
Cation
Raw
MODEL 1056
6.4.9 Slope
The display screen for Entering the conductivity/temp
Slope is shown. The default value is displayed in bold
type. Refer to the contacting conductivity Programming
flow diagram to complete this function.
6.4.10 Reference Temp
The display screen for manually entering the Reference
temperature is shown. The default value is displayed in
bold type. Refer to the contacting conductivity
Programming flow diagram to complete this function.
6.4.11 Filter
The display screen for entering the input filter value in
seconds is shown. The default value is displayed in
bold type. Refer to the contacting conductivity
Programming flow diagram to complete this function.
6.4.12 Custom Setup
The display screens for creating a custom curve for
converting conductivity to concentration is shown.
Refer to the contacting conductivity Programming
flow diagram to complete this function.
When the custom curve data entry is complete, press
ENTER. The display will confirm the determination of a
custom curve fit to the entered data by displaying this
screen:
If the custom curve fit is not completed or is
unsuccessful, the display will read as follows and the
screen will return to the beginning custom curve screen.
6.4.13 Cal Factor
Upon initial installation and power up, if 4electrode
was selected for the sensor type in the Quick Start
menus, the user enters a Cell Constant and a “Cal
Factor” using the instrument keypad. The cell constant
is needed to convert measured conductance to
conductivity as displayed on the analyzer screen. The
“Cal Factor” entry is needed increase the accuracy of
the live conductivity readings, especially at low conduc
tivity readings below 20uS/cm. Both the Cell Constant
and the “Cal Factor” are printed on the tag attached to
the 4electrode sensor/cable.
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Slope
2.00 %/ºC
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Ref Temp
(25.0ºC normal)
+25.0ºC
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Input filter
02 sec
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Custom Curve
Configure
Enter Data Points
Calculate Curve
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calculate Curve
Custom curve
fit completed.
In Process Cal
recommended.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calculate Curve
Failure
The display screen for entering Cal Factor is shown.
The default value is displayed in bold type. If necessary
after initial installation and startup, enter the “Cal
Factor” as printed on the sensor tag.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Cal Factor
0.95000/cm
47
MODEL 1056
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
6.5 TOROIDAL CONDUCTIVITY MEASUREMENT PROGRAMMING
6.5.1 Description
The section describes how to configure the Model 1056 analyzer for conductivity measurements using
inductive/toroidal sensors. The following programming and configuration functions are covered.
TABLE 64. Toroidal Conductivity Measurement Programming
Measure
Toroidal
conductivity
Sec.
Menu function:
default
6.5.2
Model:
6.5.3
Measure:
6.5.4
Range:
Auto
6.5.5
Cell K:
3.00000/cm
6.5.6
Temp Comp:
6.5.7
Slope:
6.5.8
Ref Temp:
6.5.9
Filter:
6.5.10
Custom
228
Conductivity
Slope
2.00%/°C
25.0°C
2 sec
Setup
Description
Select sensor type
Select Conductivity, Resistivity, TDS, Salinity or % conc
Select measurement Autorange or specific range
Enter the cell Constant for the sensor
Select Temp Comp: Slope, Neutral Salt, or Raw
Enter the linear temperature coefficient
Enter the Reference temp
Override the default input filter, enter 0999 seconds
Enter 25 data points in ppm and µS/cm for custom curves
A detailed flow diagram for toroidal conductivity programming is provided at the end of Sec. 6 to guide
you through all basic programming and configuration functions.
To configure the toroidal conductivity measurement
board:
1. Press MENU
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to
toroidal conductivity. Press ENTER.
The adjacent screen format will appear (factory defaults
are shown). To program any displayed function, scroll
to the desired item and press ENTER.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Configure
Model:
228
Measure:
Cond
Range:
Auto
Cell K: 3.00000/cm
RTD Offset: 0.00ºC
RTD Slope:
0
Temp Comp: Slope
Slope:
2.00%/°C
Ref Temp: 25.0°C
Filter:
2 sec
Custom Setup
The following subsections provide you with the initial display screen that appears for each configuration function.
Use the flow diagram for toroidal conductivity programming at the end of Sec. 6 and the Model 1056 live
screen prompts for each function to complete configuration and programming.
6.5.2 Sensor Model
The display screen for selecting the sensor model is
shown. The default value is displayed in bold type.
Refer to the toroidal conductivity Programming flow
diagram to complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Model
228
225
226
247
Other
48
MODEL 1056
6.5.3 Measure
The display screen for selecting the measurement is
shown. The default value is displayed in bold type.
Refer to the toroidal conductivity Programming flow diagram
to complete this function.
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Measurement
Conductivity
Resistivity
TDS
Salinity
NaOH (012%)
HCl (015%)
Low H2SO4
High H2SO4
NaCl (020%)
Custom Curve
6.5.4 Range
The display screen for Selecting Autoranging or a
specific range is shown. The default value is displayed
in bold type. Note: Ranges are shown as conductance,
not conductivity. Refer to the toroidal conductivity
Programming flow diagram to complete this function.
6.5.5 Cell Constant
The display screen for entering a cell Constant for the
sensor is shown. The default value is displayed in bold
type. Refer to the toroidal conductivity Programming
flow diagram to complete this function.
6.5.6 Temp Comp
The display screen for Selecting Temperature
Compensation as Slope, Neutral Salt, or Raw is shown.
The default value is displayed in bold type. Refer to the
toroidal conductivity Programming flow diagram to com
plete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Range
Auto
2000 mS
50 mS
2 mS
200µS
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Cell Constant
3.00000 /cm
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Temp Comp
Slope
Neutral Salt
Raw
49
MODEL 1056
6.5.7 Slope
The display screen for Entering the conductivity/temp
Slope is shown. The default value is displayed in bold
type. Refer to the toroidal conductivity Programming
flow diagram to complete this function.
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Slope
2.00%/ºC
6.5.8 Ref Temp
The display screen for manually Entering the Reference
temperature is shown. The default value is displayed in
bold type.
Refer to the toroidal conductivity
Programming flow diagram to complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
6.5.9 Filter
The display screen for entering the input filter value in
seconds is shown. The default value is displayed in
bold type.
Refer to the toroidal conductivity
Programming flow diagram to complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
6.5.10 Custom Setup
The display screens for creating custom curves for con
verting conductivity to concentration is shown. Refer to
the toroidal conductivity Programming flow diagram to
complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
When the custom curve data entry is complete, press
ENTER. The display will confirm the determination of a
custom curve fit to the entered data by displaying this
screen:
If the custom curve fit is not completed or is
unsuccessful, the display will read as follows and the
screen will return to the beginning custom curve screen.
50
SN Ref Temp
(25.0ºC normal)
+25.0ºC
SN Input filter
02 sec
SN Custom Curve
Configure
Enter Data Points
Calculate Curve
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calculate Curve
Custom curve
fit completed.
In Process Cal
recommended.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calculate Curve
Failure
MODEL 1056
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
6.6 CHLORINE MEASUREMENT PROGRAMMING
With a Chlorine measurement board installed, Model 1056 can measure any of four variants of Chlorine:
• Free Chlorine
• Total Chlorine
• Monochloramine
• pHindependent Free Chlorine
The section describes how to configure the Model 1056 analyzer for Chlorine measurements.
6.6.1 FREE CHLORINE MEASUREMENT PROGRAMMING
6.6.1.1 Description
This Chlorine subsection describes how to configure the Model 1056 analyzer for Free Chlorine measurement
using amperometric chlorine sensors. The following programming and configuration functions are covered:
TABLE 65. Free Chlorine Measurement Programming
Measure
Free
Chlorine
Sec.
Menu function:
default
Free Chlorine
Description
Select Free Chlorine, pH Ind. Free Cl. Total Cl, Monochloramine
6.6.1.2
Measure:
6.6.1.3
Units:
ppm
Select units ppm or mg/L
6.6.1.4
Filter:
5sec
Override the default input filter, enter 0999 seconds
6.6.1.5
Free Cl Correct:
6.6.1.6
Manual pH:
7.00 pH
6.6.1.7
Resolution:
0.001
Live
Select Live/Continuous pH correction or Manual
For Manual pH correction, enter the pH value
Select display resolution 0.01 or 0.001
A detailed flow diagram for programming of all chlorine measurements is provided at the end of Sec. 6 to
guide you through all basic programming and configuration functions.
To configure the chlorine measurement board for free
chlorine:
1. Press MENU
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to
chlorine. Press ENTER.
The adjacent screen format will appear (factory defaults
are shown). To program any displayed function, scroll
to the desired item and press ENTER.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Configure
Measure: Free Chlorine
Units:
ppm
Filter:
5sec
Free Cl Correct:
Live
Manual pH:
7.00 pH
Resolution:
0.001
51
MODEL 1056
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
The following subsections provide you with the initial display screen that appears for each configuration function.
Use the flow diagram for chlorine programming at the end of Sec. 6 and the Model 1056 live screen prompts
for each function to complete configuration and programming.
6.6.1.2 Measure
The display screen for selecting the measurement is
shown. The default value is displayed in bold type.
Refer to the Chlorine Programming flow diagram to
complete this function.
6.6.1.3 Units
The display screen for selecting units as ppm or mg/L
is shown. The default value is displayed in bold type.
Refer to the Chlorine Programming flow diagram to
complete this function.
6.6.1.4 Filter
The display screen for entering the input filter value in
seconds is shown. The default value is displayed in
bold type. Refer to the Chlorine Programming flow
diagram to complete this function.
6.6.1.5 Free Chlorine pH Correction
The display screen for Selecting Live/Continuous pH
correction or Manual pH correction is shown. The
default value is displayed in bold type. Refer to the
Chlorine Programming flow diagram to complete this
function.
6.6.1.6 Manual pH Correction
The display screen for manually entering the pH value
of the measured process liquid is shown. The default
value is displayed in bold type. Refer to the Chlorine
Programming flow diagram to complete this function.
6.6.1.7 Resolution
The display screen for selecting display resolution as
0.001 or 0.01 is shown. The default value is displayed
in bold type. Refer to the Chlorine Programming flow
diagram to complete this function.
52
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Measurement
Free Chlorine
pH Independ. Free Cl
Total Chlorine
Monochloramine
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Units
ppm
mg/L
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Input filter
05 sec
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Free Cl
pH Correction
Live/Continuous
Manual
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Manual pH
07.00 pH
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Resolution
0.001
0.01
MODEL 1056
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
6.6.2 TOTAL CHLORINE MEASUREMENT PROGRAMMING
6.6.2.1 Description
This Chlorine subsection describes how to configure the Model 1056 analyzer for Total Chlorine measurement using
amperometric chlorine sensors. The following programming and configuration functions are covered:
TABLE 66. Total Chlorine Measurement Programming
Measure
Total
Chlorine
Sec.
Menu function:
default
6.6.2.2
Measure:
6.6.2.3
Units:
ppm
6.6.2.4
Filter:
5sec
6.6.2.5
Resolution:
0.001
Free Chlorine
Description
Select Free Chlorine, pH Ind. Free Cl. Total Cl, Monochloramine
Select units ppm or mg/L
Override the default input filter, enter 0999 seconds
Select 0.01 or 0.001 display resolution
A detailed flow diagram for programming of all chlorine measurements is provided at the end of Sec. 6 to
guide you through all basic programming and configuration functions.
To configure the chlorine measurement board for total
chlorine:
1. Press MENU
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to
chlorine. Press ENTER.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Configure
Measure: Free Chlorine
Units:
ppm
Filter:
5sec
Resolution:
0.001
The adjacent screen format will appear (factory defaults
are shown). To program any displayed function, scroll
to the desired item and press ENTER.
The following subsections provide you with the initial display screen that appears for each configuration function.
Use the flow diagram for chlorine programming at the end of Sec. 6 and the Model 1056 live screen prompts for
each function to complete configuration and programming.
6.6.2.2 Measure
The display screen for selecting the measurement is
shown. The default value is displayed in bold type.
Refer to the chlorine Programming flow diagram to
complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
6.6.2.3 Units
The display screen for selecting units as ppm or mg/L
is shown. The default value is displayed in bold type.
Refer to the Chlorine Programming flow diagram to
complete this function
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Measurement
Free Chlorine
pH Independ. Free Cl
Total Chlorine
Monochloramine
SN Units
ppm
mg/L
53
MODEL 1056
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
6.6.2.4 Filter
The display screen for entering the input filter value in
seconds is shown. The default value is displayed in
bold type. Refer to the Chlorine Programming flow
diagram to complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Input filter
05 sec
6.6.2.5 Resolution
The display screen for selecting display resolution as
0.001 or 0.01 is shown. The default value is displayed
in bold type. Refer to the Chlorine Programming flow
diagram to complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Resolution
0.001
0.01
6.6.3 MONOCHLORAMINE MEASUREMENT PROGRAMMING
6.6.3.1 Description
This Chlorine subsection describes how to configure the Model 1056 analyzer for Monochloramine measurement
using amperometric chlorine sensors. The following programming and configuration functions are covered:
TABLE 67. Monochloramine Measurement Programming
Measure
Sec.
Monochloramine 6.6.3.2
6.6.3.3
6.6.3.4
6.6.3.5
Menu function:
default
Measure: Free Chlorine
Units:
ppm
Filter:
5sec
Resolution:
0.001
Description
Select Free Chlorine, pH Ind. Free Cl. Total Cl, Monochloramine
Select units ppm or mg/L
Override the default input filter, enter 0999 seconds
Select 0.01pH or 0.1ppm/mg/L for display Resolution
A detailed flow diagram for programming of all chlorine measurements is provided at the end of Sec. 6 to
guide you through all basic programming and configuration functions.
To configure the chlorine measurement board for
monochloramine:
1. Press MENU
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to
chlorine. Press ENTER.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Configure
Measure: Free Chlorine
Units:
ppm
5sec
Filter:
Resolution:
0.001
The following screen format will appear (factory defaults
are shown). To program any displayed function, scroll
to the desired item and press ENTER.
The following subsections provide you with the initial display screen that appears for each configuration function.
Use the flow diagram for chlorine programming at the end of Sec. 6 and the Model 1056 live screen prompts
for each function to complete configuration and programming.
6.6.3.2Measure: Monochloramine
The display screen for selecting the measurement is
shown. The default value is displayed in bold type.
Refer to the Chlorine Programming flow diagram to
complete this function.
54
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Measurement
Free Chlorine
pH Independ. Free Cl
Total Chlorine
Monochloramine
MODEL 1056
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
6.6.3.3 Units
The display screen for selecting units as ppm or mg/L is
shown. The default value is displayed in bold type.
Refer to the Chlorine Programming flow diagram to
complete this function.
6.6.3.4 Filter
The display screen for entering the input filter value in
seconds is shown. The default value is displayed in
bold type. Refer to the Chlorine Programming flow diagram
to complete this function.
6.6.3.5 Resolution
The display screen for selecting display resolution as
0.001 or 0.01 is shown. The default value is displayed
in bold type. Refer to the Chlorine Programming flow
diagram to complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Units
ppm
mg/L
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Input filter
05 sec
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Resolution
0.001
0.01
6.6.4 pHINDEPENDENT FREE CHLORINE MEASUREMENT PROGRAMMING
6.6.4.1 Description
This Chlorine subsection describes how to configure the Model 1056 analyzer for Free Chlorine measurements
using the pHindependent free chlorine sensor, Model 498CL01, manufactured by Rosemount Analytical. The following
programming and configuration functions are covered:
TABLE 68. pHindependent Free Chlorine Measurement Programming
Measure
pHindependent
Free Chlorine
Sec.
6.6.4.2
Menu function:
default
Measure: pH Indep Free Cl
6.6.4.3
Units:
ppm
6.6.4.4
Filter:
5sec
6.6.4.5
Resolution:
0.001
Description
Select Free Chlorine, pH Ind. Free Cl. Total Cl, Monochloramine
Select units ppm or mg/L
Override the default input filter, enter 0999 seconds
Select 0.01pH or 0.1ppm/mg/L for display Resolution
A detailed flow diagram for programming of all chlorine measurements is provided at the end of Sec. 6 to
guide you through all basic programming and configuration functions.
To configure the chlorine measurement board for
pHindependent free chlorine:
1. Press MENU
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to
chlorine. Press ENTER.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Configure
Measure: Free Chlorine
Units:
ppm
Filter:
5sec
Resolution:
0.001
The adjacent screen format will appear (factory defaults
are shown). To program any displayed function, scroll
to the desired item and press ENTER.
55
MODEL 1056
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
The following subsections provide you with the initial display screen that appears for each configuration function.
Use the flow diagram for chlorine programming at the end of Sec. 6 and the Model 1056 live screen prompts
for each function to complete configuration and programming.
6.6.4.2 Measurement: pHindependent Free Chlorine
The display screen for selecting the measurement is
shown. The default value is displayed in bold type.
Refer to the chlorine Programming flow diagram to
complete this function.
6.6.4.3 Units
The display screen for selecting units as ppm or mg/L
is shown. The default value is displayed in bold type.
Refer to the Chlorine Programming flow diagram to
complete this function.
6.6.4.4 Filter
The display screen for entering the input filter value in
seconds is shown. The default value is displayed in
bold type. Refer to the Chlorine Programming flow
diagram to complete this function.
6.6.4.5 Resolution
The display screen for selecting display resolution as
0.001 or 0.01 is shown. The default value is displayed
in bold type. Refer to the Chlorine Programming flow
diagram to complete this function.
56
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Measurement
Free Chlorine
pH Independ. Free Cl
Total Chlorine
Monochloramine
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Units
ppm
mg/L
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Input filter
05 sec
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Resolution
0.001
0.01
MODEL 1056
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
6.7 OXYGEN MEASUREMENT PROGRAMMING
6.7.1 Description
This section describes how to configure the Model 1056 analyzer for dissolved and gaseous oxygen measurement
using amperometric oxygen sensors. The following programming and configuration functions are covered:
TABLE 69. Oxygen Measurement Programming
Measure
Oxygen
Sec.
Menu function:
default
Description
6.7.2
Type:
Water/Waste
Select Water/Waste, Trace. BioRx, BioRxOther, Brew, %O2 In Gas
6.7.3
Units:
ppm
Select ppm, mg/L, ppb, µg/L, % Sat, %O2Gas, ppm OxygenGas
6.7.4
Partial Press:
mmHg
Select mm Hg, in Hg. atm, kPa, mbar or bar for Partial pressure
6.7.5
Salinity:
00.0‰
Enter Salinity as ‰
6.7.6
Filter:
6.7.7
Pressure Units:
6.7.8
Use Press:
5sec
Override the default input filter, enter 0999 seconds
bar
Select pressure units: mm Hg, in Hg,. Atm, kPa, mbar, bar
At Air Cal
Select atmospheric pressure source – internal or mA Input
A detailed flow diagram for oxygen programming is provided at the end of Sec. 6 to guide you through all basic pro
gramming and configuration functions.
To configure the Oxygen measurement board:
1. Press MENU
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to Oxygen.
Press ENTER.
The adjacent screen format will appear (factory defaults are
shown). To program any displayed function, scroll to the
desired item and press ENTER.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Configure
Type:
Water/Waste
Units:
ppm
Partial Press: mmHg
Salinity:
00.0‰
Filter:
5sec
Pressure Units: bar
Use Press: At Air Cal
Custom Setup
The following subsections show the initial display screen that appears for each configuration function. Use the
flow diagram for oxygen programming at the end of Sec. 6 and the Model 1056 live screen prompts for each
function to complete configuration and programming.
6.7.2 Oxygen Measurement application
The display screen for programming the measurement
is shown. The default value is displayed in bold type.
Refer to the Oxygen Programming flow diagram to
complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Type
Water/Waste
Trace Oxygen
BioRxRosemount
BioRxOther
Brewing
Oxygen In Gas
6.7.3 Units
The display screen for selecting units as ppm , mg/L,
ppb, µg/L, % Saturation, %Oxygen in Gas, or ppm
Oxygen in Gas is shown. The default value is displayed
in bold type. Refer to the Oxygen Programming flow
diagram to complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Units
ppm
mg/L
ppb
µg/L
% Saturation
Partial Pressure
% Oxygen In Gas
ppm Oxygen In Gas
57
MODEL 1056
6.7.4 Partial Press
The display screen for selecting pressure units for
Partial pressure is shown. This selection is needed if
the specified measurement is Partial pressure. The
default value is displayed in bold type. Refer to the
Oxygen Programming flow diagram to complete this
function.
6.7.5 Salinity
The display screen for Entering the Salinity (as parts
per thousand) of the process liquid to be measured is
shown. The default value is displayed in bold type.
Refer to the Oxygen Programming flow diagram to
complete this function.
Enter Salinity as ‰
6.7.6 Filter
The display screen for entering the input filter value in
seconds is shown. The default value is displayed in
bold type. Refer to the Oxygen Programming flow
diagram to complete this function.
6.7.7 Pressure Units
The display screen for selecting pressure units for
atmospheric pressure is shown. This selection is needed
for the display of atmospheric pressure measured by
the onboard pressure transducer on the Oxygen
measurement board. The default value is displayed in
bold type. Refer to the Oxygen Programming flow
diagram to complete this function.
6.7.8 Use Pressure
The display screen for selecting atmospheric pressure
source. The default value is displayed in bold type.
Refer to the Oxygen Programming flow diagram to com
plete this function.
58
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Partial Press
mm Hg
in Hg
atm
kPa
mbar
bar
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Salinity
00.0 ‰
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Input filter
05 sec
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
Pressure Units
mm Hg
in Hg
atm
kPa
mbar
bar
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Use Pressure?
At Air Cal
mA Input
MODEL 1056
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
6.8 OZONE MEASUREMENT PROGRAMMING
6.8.1 Description
This section describes how to configure the 1056 analyzer for ozone measurement using amperometric ozone sen
sors. The following programming and configuration functions are covered:
TABLE 610. Ozone Measurement Programming
Measure
Ozone
Sec.
Menu function:
default
6.8.2
Units:
ppm
6.8.3
Filter:
5sec
6.8.4
Resolution:
0.001
Description
Select units ppm, mg/L, ppb, µg/L
Override the default input filter, enter 0999 seconds
Select 0.01or 0.001 for display resolution
A detailed flow diagram for ozone programming is provided at the end of Sec. 6 to guide you through all
basic programming and configuration functions.
To configure the Ozone measurement board:
1. Press MENU
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to
Ozone. Press ENTER.
The adjacent screen format will appear (factory defaults
are shown). To program any displayed function, scroll to
the desired item and press ENTER.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Configure
Units:
ppm
Filter:
5 sec
Resolution: 0.001
The following subsections show the initial display screen that appears for each configuration function. Use the
flow diagram for ozone programming at the end of Sec. 6 and the Model 1056 live screen prompts for each
function to complete configuration and programming.
Note: Ozone measurement boards are detected automatically by the analyzer. No measurement selection is necessary.
6.8.2 Units
The display screen for selecting measurement units is
shown. The default value is displayed in bold type.
Refer to the Ozone Programming flow diagram to
complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
6.8.3 Filter
The display screen for entering the input filter value in
seconds is shown. The default value is displayed in
bold type. Refer to the Ozone Programming flow
diagram to complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
6.8.4 Resolution
The display screen for selecting display resolution as
0.001 or 0.01 is shown. The default value is displayed
in bold type. Refer to the Ozone Programming flow
diagram to complete this function.
SN Units
ppm
mg/L
ppb
µg/L
SN Input filter
05 sec
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Resolution
0.001
0.01
59
MODEL 1056
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
6.9 TURBIDITY MEASUREMENT PROGRAMMING
6.9.1 DESCRIPTION
This section describes how to configure the Model 1056 analyzer for Turbidity measurements. The following
programming and configuration functions are covered.
TABLE 611 TURBIDITY MEASUREMENT PROGRAMMING
Measure
Turbidity
Sec.
Menu function:
6.9.2
Measurement type:
Turbidity
6.9.3
Measurement units:
NTU
6.9.4
Enter TSS* Data:
6.9.5
Filter:
6.9.6
Bubble Rejection:
default
Description
Select Turbidity or TSS calculation (estimated TSS)
NTU, FTU, FNU
Enter TSS and NTU data to calculate TSS based on Turbidity
20 sec
On
Override the default input filter, enter 0999 seconds
Intelligent software algorithm to eliminate erroneous readings
caused by bubble accumulation in the sample
*TSS: Total Suspended Solids
A detailed flow diagram for Turbidity programming is provided at the end of Sec. 6 to guide you through all basic
programming and configuration functions.
To
1.
2.
3.
4.
configure the Turbidity measurement board:
Press MENU
Scroll down to Program. Press ENTER.
Scroll down to Measurement. Press ENTER.
Select Sensor 1 or Sensor 2 corresponding to Turbidity. Press ENTER.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Configure
Measure:
Turbidity
Units:
NTU
Enter TSS Data
Filter:
20sec
Bubble Rejection: On
The following screen format will appear (factory defaults are shown).
To program Turbidity, scroll to the desired item and press ENTER.
The following subsections provide you with the initial display screen that appears for each programming routine.
Use the flow diagram for Turbidity programming at the end of Sec. 6 and the live screen prompts to complete
programming.
6.9.2 Measurement
The display screen for selecting the measurement is
shown. The default measurement is displayed in bold
type. Refer to the Turbidity Programming flow diagram
to complete this function.
6.9.3 Units
The display screen for selecting the measurement
units is shown. The default value is displayed in bold
type. Refer to the Turbidity Programming flow diagram
to complete this function.
60
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Measurement
Turbidity
Calculated TSS
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Units
NTU
FTU
FNU
MODEL 1056
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
If TSS data (Total Suspended Solids) calculation is
selected, the following screen will be displayed. Refer
to the Turbidity programming flow diagram to complete
this function.
6.9.4 Enter TSS Data
The display screen for entering TSS Data is shown.
The default values are displayed. Refer to the
Turbidity Programming flow diagram to complete this
function
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Units
ppm
mg/L
none
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN TSS Data
Pt1 TSS:
0.000ppm
Pt1 Turbid: 0.000NTU
Pt2 TSS:
100.0ppm
Pt2 Turbid: 100.0NTU
Calculate
Note: Based on userentered NTU data, calculating
TSS as a straight line curve could cause TSS to go
below zero. The following screen lets users know that
TSS will become zero below a certain NTU value.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN TSS Data
Calculation Complete
Calculated TSS = 0 below
xxxx NTU
The following illustration shows the potential for calculated TSS to go below zero
Normal case: TSS is always a positive number when Turbidity is a positive number.
Abnormal case: TSS can be a negative number when Turbidity is a positive number.
61
MODEL 1056
When the TSS data entry is complete, press ENTER.
The display will confirm the determination of a TSS
straight line curve fit to the entered NTU/turbidity data
by displaying this screen:
The following screen may appear if TSS calculation is
unsuccessful. Reentry of NTU and TSS data is
required.
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN TSS Data
Calculation
Complete
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN TSS Data
Data Entry Error
Press EXIT
6.9.5 Filter
The display screen for entering the input filter value in
seconds is shown. The default value is displayed in
bold type. Refer to the Turbidity Programming flow
diagram to complete this function.
6.9.6 Bubble Rejection
Bubble rejection is an internal software algorithm that
characterizes turbidity readings as bubbles as
opposed to true turbidity of the sample. With Bubble
rejection enabled, these erroneous readings are elimi
nated from the live measurements shown on the dis
play and transmitted via the current outputs.
The display screen for selecting bubble rejection algo
rithm is shown. The default setting is displayed in bold
type. Refer to the Turbidity Programming flow diagram
to complete this function.
62
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Input Filter
020sec
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Bubble Rejection
On
Off
MODEL 1056
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
6.10 FLOW MEASUREMENT PROGRAMMING
6.10.1 DESCRIPTION
This section describes how to configure the 1056 analyzer for flow measurement when used with a compatible pulse
flow sensor. The following programming and configuration functions are covered.
TABLE 612 FLOW MEASUREMENT PROGRAMMING
Measure
Flow
To
1.
2.
3.
4.
Sec.
Menu function:
6.10.2
Measurement type
6.10.3
Measurement units:
6.10.4
Enter TSS* Data:
default
Pulse Flow
GPH
0 Sec
Description
Select Pulse Flow or mA Current Input
Select GPM, GPH, cu ft/min, cu ft/hour, LPM, L/hour, m3/hr
Override the default input filter, enter 0999 seconds
configure the flow measurement board:
Press MENU
Scroll down to Program. Press ENTER.
Scroll down to Measurement. Press ENTER.
Select Sensor 1 or Sensor 2 corresponding to
flow. Press ENTER.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Configure
Measure: Pulse Flow
Units:
GPM
Filter:
5sec
The following screen format will appear (factory defaults
are shown).
To program pulse flow, scroll to the desired item and press ENTER.
The following subsections provide you with the initial display screen that appears for each programming routine.
Use the diagram for pulse flow programming at the end of Sec. 6 and the live screen prompts to complete
programming.
6.10.2 Measurement
The display screen for selecting the measurement is
shown. The default measurement is displayed in bold
type. Refer to the pulse flow Programming diagram to
complete this function.
6.10.3 Units
The display screen for selecting measurement units is
shown. The default units are displayed in bold type.
Refer to the pulse flow Programming diagram to com
plete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Measurement
Pulse Flow
mA Input
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Units
GPM
GPH
cu ft/min
cu ft/hour
L/min
L/hour
m3/hour
6.10.4 Filter
The display screen for entering the input filter value in
seconds is shown. The default value is displayed in
bold type. Refer to the pulse flow Programming dia
gram to complete this function.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Input Filter
005sec
63
MODEL 1056
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
6.11 CURRENT INPUT PROGRAMMING
6.11.1 DESCRIPTION
This section describes how to configure the Model 1056 analyzer for current input measurement when wired to an
external device that transmits 420mA or 020mA analog current output. The following programming and configu
ration functions are covered.
TABLE 613 CURRENT INPUT PROGRAMMING
Measure
Current
Input
Sec.
Menu function:
6.11.2
Measurement type
6.11.3
mA Input
default
mA input
Temperature
ºC
Description
Override the default (Flow) and select mA current input
Select Temperature, Pressure, Flow or Other
6.11.4
Measurement units:
6.11.5
Input Range:
420mA
Select 420mA or 020mA
6.11.6
Low Value:
0.000ºC
Enter the low measurement value to assign to 4mA
6.11.7
High Value:
100.0ºC
Enter the high measurement value to assign to 20mA
6.11.8
Filter:
05 sec
Override the default input filter, enter 0999 seconds
Select measurement units based on selected input device type
A detailed flow diagram for current input programming is provided at the end of Sec. 6 to guide you through all
basic programming and configuration functions.
To
1.
2.
3.
4.
configure the current input measurement board:
Press MENU
Scroll down to Program. Press ENTER.
Scroll down to Measurement. Press ENTER.
Select Sensor 1 or Sensor 2 corresponding to
current input. Press ENTER.
Note that factory default is Pulse Flow not mA Input.
The user must override the factory default and select
mA Input to enable the current input functionality. Upon
selecting mA Input, the following menu screen will
appear to allow complete programming of mA Current
Input.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Configure
Measure:
mA Input
mA Input: Temperature
Units:
ºC
Input Range:
420mA
Low Value:
High Value :
Filter:
0.001%
100.0%
5sec
To program current input, scroll to the desired item and press ENTER.
The following subsections provide you with the initial display screen that appears for each programming routine.
Use the flow diagram for current input programming at the end of Sec. 6 and the live screen prompts to complete
programming.
6.11.2 Measurement
The display screen for selecting the signal board func
tionality is shown. The default value is displayed in
bold type. Scroll down to select mA Input to enable
the current input functionality. Refer to the current
input Programming flow diagram to complete this func
tion.
6.11.3 mA Input
The display screen for selecting the type of measure
ment is shown. The default measurement type for mA
Input is displayed in bold type. Refer to the current
input Programming flow diagram to complete this function.
64
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Measurement
Pulse Flow
mA Input
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN mA Input
Temperature
Pressure
Flow
Other
MODEL 1056
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
6.11.4 Units
The display screen for selecting measurement units is
shown. The default value for temperature is displayed in
bold type. Refer to the current input Programming flow
diagram to complete this function.
If Pressure is selected as the measurement type for
mA Input, the following display screen is shown:
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Units
°C
ºF
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Units
mm Hg
in Hg
atm
kPa
mbar
bar
The current input board can also be used to accept a
420mA current input from a pulse flow sensor. If
Flow is selected as the measurement type for the
420mA current input board, the following display
screen is shown:
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Units
GPM
GPH
cu ft/min
cu ft/hour
L/min
L/hour
m3/hour
Current input can serve as a universal measurement
board. 420mA current input can be accepted from any
device and assigned to represent a wide range of
measurements. If Other is selected as the measure
ment type for the 420mA current input board, the fol
lowing display screen is shown:
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Units
%
% Saturation
pH
mV
Any of the following units can also be selected to represent the 420mA current input. Simply scroll down to
identify and select the desired measurement units as listed in the table below.
µS/cm
ppm
µg/L
NTU
ft/sec
mS/cm
ppb
m/sec
mg/L
FTU
MΩcm
g/L
FNU
kΩcm
‰
6.11.5 Input Range
The display screen for selecting the Input Range is
shown. The default value for mA Input is displayed in
bold type. Refer to the current input Programming
flow diagram to complete this function.
none
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Input Range
420mA
020mA
65
MODEL 1056
6.11.6 Low Value
The display screen for entering the Low Value to be
assigned to 4mA (or 0mA) current input is shown. The
default value for temperature is displayed in bold type.
Refer to the current input Programming flow diagram
to complete this function.
6.11.7 High Value
The display screen for entering the High Value to be
assigned to 20mA current input is shown. The default
value for temperature is displayed in bold type. Refer
to the current input Programming flow diagram to com
plete this function.
6.11.8 Filter
The display screen for entering the input filter value in
seconds is shown. The default value is displayed in
bold type. Refer to the current input Programming
diagram to complete this function. .
66
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Low Value
0.000ºC
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN High Value
100.0ºC
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Input Filter
005sec
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
FIGURE 61 Configuring pH/ORP Measurements
MODEL 1056
67
FIGURE 62 Configure Contacting Measurements
MODEL 1056
68
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
FIGURE 63 Configure Toroidal Measurements
MODEL 1056
69
FIGURE 65 Configure Oxygen Measurements
MODEL 1056
70
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
MODEL 1056
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
FIGURE 64 Configure Chlorine Measurements
FIGURE 66 Configure Ozone Measurements
71
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
FIGURE 67 Configure Turbidity Measurement
MODEL 1056
72
MODEL 1056
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
FIGURE 68 Configure Flow Measurement
FIGURE 69 Configure mA Current Input Measurement
73
MODEL 1056
74
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
This page left blank intentionally
MODEL 1056
SECTION 7.0
CALIBRATION
SECTION 7.0
CALIBRATION
7.1
7.2
7.3
7.4
7.5
7.6
CALIBRATION – INTRODUCTION
pH CALIBRATION
ORP CALIBRATION
CONTACTING CONDUCTIVITY CALIBRATION
TOROIDAL CONDUCTIVITY CALIBRATION
CHLORINE CALIBRATION
7.6.1 FREE CHLORINE
7.6.2 TOTAL CHLORINE
7.6.3 MONOCHLORAMINE
7.6.4 pHINDEPENDENT FREE CHLORINE
7.7 OXYGEN CALIBRATION
7.8 OZONE CALIBRATION
7.9 TEMPERATURE CALIBRATION
7.10 TURBIDITY CALIBRATION
7.11 FLOW CALIBRATION
7.1 CALIBRATION – INTRODUCTION
Calibration is the process of adjusting or standardizing
the analyzer to a lab test or a calibrated laboratory
instrument, or standardizing to some known reference
(such as a commercial buffer).
The autorecognition feature of the analyzer will enable
the appropriate calibration screens to allow calibration for
any single sensor configuration or dual sensor config
uration of the analyzer. Completion of Quick Start upon
first power up enables live measurements but does not
ensure accurate readings in the lab or in process.
Calibration should be performed with each attached
sensor to ensure accurate, repeatable readings.
This section covers the following programming and
configuration functions:
4. Standardization calibration (1point) for pH, ORP
and Redox (pH Cal Sec.7.2 and 7.3)
5. Entering the cell constant of a conductivity
sensor (Conductivity Cal Sec. 7.4 and 7.5)
6. Calibrating the sensor in a conductivity standard
Conductivity Cal Sec. 7.4 and 7.5)
7. Calibrating the analyzer to a laboratory
instrument (Contacting Conductivity Cal Sec.7.4)
8. Zeroing an chlorine, oxygen or ozone sensor
(Amperometric Cal Sec’s 7.6, 7.7, 7.8)
9. Calibrating an oxygen sensor in air (Oxygen Cal
Sec’s 7.6)
10. Calibrating the sensor to a sample of known
concentration (Amperometric Cal Sec’s 7.6, 7.7, 7.8)
11. Enter a manual reference temperature for
temperature compensation of the process
measurement
1. Auto buffer cal for pH (pH Cal Sec.7.2)
2. Manual buffer cal for pH (pH Cal Sec.7.2)
3. Set calibration stabilization criteria for pH (pH Cal
Sec.7.2)
75
MODEL 1056
SECTION 7.0
CALIBRATION
7.2 pH CALIBRATION
7.2.1 DESCRIPTION
New sensors must be calibrated before use. Regular recalibration is also necessary. Use auto calibration instead of manual
calibration. Auto calibration avoids common pitfalls and reduces errors. The analyzer recognizes the buffers and uses
temperaturecorrected pH values in the calibration. Once the Model 1056 successfully completes the calibration, it calcu
lates and displays the calibration slope and offset. The slope is reported as the slope at 25°C.
THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1056 WITH A pH SENSOR. THE FOLLOWING
CALIBRATION ROUTINES ARE COVERED.
TABLE 71
Measure Sec.
pH
7.2.2
pH Calibration Routines
Menu function:
default
Description
Auto Calibration
pH
2 point buffer calibration with auto buffer recognition
7.2.3
Manual Calibration
pH
2 point buffer calibration with manual buffer value entry
7.2.4
Entering A Known Slope Value pH
Slope calibration with manual entry of known slope value
7.2.5
Standardization
1 point buffer calibration with manual buffer value entry
pH
A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines.
To calibrate pH:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to
pH. Press ENTER.
4. Select pH. Press ENTER.
The following screen will appear. To calibrate pH or
Temperature scroll to the desired item and press
ENTER.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibrate?
pH
Temperature
The following subsections show the initial display screen that appears for each calibration routine. Use the flow
diagram for pH calibration at the end of Sec. 7 and the live screen prompts to complete calibration.
7.2.2 AUTO CALIBRATION — pH
This screen appears after selecting pH calibration.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN pH Cal
Buffer Cal
Standardize
Slope:
59.16mV/pH
Offset:
600 mV
Note that pH auto calibration criteria can be changed.
The following criteria can be adjusted:
Stabilization time (default 10 sec.)
Stabilization pH value (default 0.02 pH)
Type of Buffer used for AUTO CALIBRATION
(default is Standard, noncommercial buffers).
The following commercial buffer tables are recognized
by the analyzer:
Standard (NIST plus pH7)
DIN 19267
Ingold
Merck
76
The following screen will appear to allow adjustment of
these criteria:
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Setup
Stable Time:
10 sec
Stable Delta: 0.02 pH
Buffer:
Standard
MODEL 1056
SECTION 7.0
CALIBRATION
The following screen will appear if the auto cal is
successful. The screen will return to the pH Buffer Cal
Menu.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN pH Auto Cal
Slope: 59.16 mV/pH
Offset:
60 mV
The following screens may appear if the auto cal is unsuccessful.
1. A High Slope Error will generate this screen display:
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN pH Auto Cal
High Slope Error
Calculated: 62.11 mV/pH
Max: 62.00 mV/pH
Press EXIT
2. A Low Slope Error will generate this screen display:
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN pH Auto Cal
Low Slope Error
Calculated: 39.11mV/pH
Min: 40.00 mV/pH
Press EXIT
3. An Offset Error will generate this screen display:
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN pH Auto Cal
Offset Error
Calculated: 61.22mV
Max:
60.00mV
Press EXIT
7.2.3 MANUAL CALIBRATION — pH
New sensors must be calibrated before use. Regular
recalibration is also necessary. Use manual calibration
if nonstandard buffers are being used; otherwise, use
auto calibration. Auto calibration avoids common pitfalls
and reduces errors.
The adjacent appears after selecting Manual pH calibra
7.2.4 ENTERING A KNOWN SLOPE VALUE — pH
If the electrode slope is known from other measure
ments, it can be entered directly in the Model 1056 ana
lyzer. The slope must be entered as the slope at 25°C.
7.2.5 STANDARDIZATION — pH
The pH measured by the Model 1056 analyzer can be
changed to match the reading from a second or referee
instrument. The process of making the two readings
agree is called standardization. During standardization,
the difference between the two pH values is converted
to the equivalent voltage. The voltage, called the reference
offset, is added to all subsequent measured cell voltages
before they are converted to pH. If a standardized sensor
is placed in a buffer solution, the measured pH will differ
from the buffer pH by an amount equivalent to the
standardization offset.
tion.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN pH Manual Cal
Buffer 1
Buffer 2
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN pH Slope@25ºC
59.16 mV/pH
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Enter Value
07.00pH
77
MODEL 1056
SECTION 7.0
CALIBRATION
The following screen may appear if ORP Cal is unsuccessful.
An Offset Error will generate this screen display:
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Standardize
Offset Error
Calculated: 96mV
Max:
60mV
Press EXIT
If the ORP Cal is successful, the screen will return to the
Cal submenu.
7.3 ORP CALIBRATION
7.3.1 DESCRIPTION
For process control, it is often important to make the measured ORP agree with the ORP of a standard solution. During
calibration, the measured ORP is made equal to the ORP of a standard solution at a single point.
THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1056 WITH AN ORP SENSOR. THE FOL
LOWING CALIBRATION ROUTINE IS COVERED.
TABLE 72
Measure
ORP
ORP Calibration Routine
Sec.
7.3.2
Menu function: default
Description
Standardization — ORP 1 point buffer calibration with manual buffer value entry
A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines.
To calibrate ORP:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to
ORP. Press ENTER.
4. Select ORP. Press ENTER.
The following screen will appear. To calibrate ORP or
Temperature, scroll to the desired item and press
ENTER.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibrate?
ORP
Temperature
The following subsections show the initial display screen that appears for each calibration routine. Use the flow
diagram for ORP calibration at the end of Sec. 7 and the live screen prompts to complete calibration.
7.3.2 STANDARDIZATION — ORP
For process control, it is often important to make the
measured ORP agree with the ORP of a standard
solution. During calibration, the measured ORP is made
equal to the ORP of a standard solution at a single
point. This screen appears after selecting ORP cali
bration:
Cal submenu.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Enter Value
+0600 mV
If the ORP Cal is successful, the screen will return to the
The following screen may appear if ORP Cal is
unsuccessful.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Standardize
Offset Error
Calculated: 61.22mV
Max:
60.00mV
Press EXIT
78
MODEL 1056
SECTION 7.0
CALIBRATION
7.4 CONTACTING CONDUCTIVITY CALIBRATION
7.4.1 DESCRIPTION
PLACING A NEW CONDUCTIVITY SENSOR IN SERVICE
New conductivity sensors rarely need calibration. The cell
constant printed on the label is sufficiently accurate for most
applications.
CALIBRATING AN INSERVICE CONDUCTIVITY SENSOR
1. After a conductivity sensor has been in service for a period
of time, recalibration may be necessary. There are three
ways to calibrate a sensor.
a. Use a standard instrument and sensor to measure the
conductivity of the process stream. It is not necessary to
remove the sensor from the process piping. The temperature
correction used by the standard instrument may not exactly
match the temperature correction used by the Model 1056.
To avoid errors, turn off temperature correction in both the
analyzer and the standard instrument.
b. Place the sensor in a solution of known conductivity and
make the analyzer reading match the conductivity of the
standard solution. Use this method if the sensor can be easily
removed from the process piping and a standard is
available. Be careful using standard solutions having
conductivity less than 100 µS/cm. Low conductivity standards
are highly susceptible to atmospheric contamination. Avoid
calibrating sensors with 0.01/cm cell constants against
conductivity standards having conductivity greater than 100
µS/cm. The resistance of these solutions may be too low for
an accurate measurement. Calibrate sensors with 0.01/cm
cell constant using method c.
c. To calibrate a 0.01/cm sensor, check it against a standard
instrument and 0.01/cm sensor while both sensors are
measuring water having a conductivity between 5 and 10
µS/cm. To avoid drift caused by absorption of atmospheric
carbon dioxide, saturate the sample with air before making
the measurements.
To ensure adequate flow past the sensor during calibration,
take the sample downstream from the sensor. For best
results, use a flowthrough standard cell. If the process
temperature is much different from ambient, keep
connecting lines short and insulate the flow cell.
THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1056 WITH AN ATTACHED CONTACTING
CONDUCTIVITY SENSOR. THE FOLLOWING CALIBRATION ROUTINES ARE COVERED.
TABLE 73
Contacting Conductivity Calibration Routines
Menu function:
default
Description
Measure
Sec.
Contacting
Conductivity
7.4.2
Cell K:
7.4.3
Zero Cal
Zero the analyzer with the sensor attached
7.4.4
In Process Cal
Standardize the sensor to a known conductivity
7.4.5
Meter Cal
Calibrate the analyzer to a lab conductivity instrument
7.4.6
Cal Factor:
1.00000/cm
0.95000/cm
Enter the cell Constant for the sensor
Enter the Cal Factor for 4Electrode sensors from the sensor tag
A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines
To calibrate contacting conductivity:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
S1: 1.234µS/cm 25.0ºC
3. Select Sensor 1 or Sensor 2 corresponding to
S2: 12.34pH
25.0ºC
contacting conductivity. Press ENTER.
SN Calibrate?
4. Select Conductivity. Press ENTER.
The adjacent screen will appear.
To calibrate
Conductivity or Temperature, scroll to the desired item
and press ENTER.
The following subsections show the initial display
screen that appears for each calibration routine. Use
the flow diagram for Conductivity calibration at the
end of Sec. 7 and the live screen prompts for each rou
tine to complete calibration.
The adjacent screen appears
Conductivity calibration:
after
selecting
Conductivity
Temperature
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibration
Zero Cal
In Process Cal
Meter Cal
Cell K: 1.00000/cm
79
MODEL 1056
7.4.2 ENTERING THE CELL CONSTANT
New conductivity sensors rarely need calibration. The
cell constant printed on the label is sufficiently accurate
for most applications. The cell constant should be
entered:
• When the unit is installed for the first time
• When the probe is replaced
The display screen for entering a cell Constant for the
sensor is shown. The default value is displayed in bold
type.
7.4.3 ZEROING THE INSTRUMENT
This procedure is used to compensate for small offsets
to the conductivity signal that are present even when
there is no conductivity to be measured. This procedure
is affected by the length of extension cable and should
always be repeated if any changes in extension cable or
sensor have been made. Electrically connect the
conductivity probe as it will actually be used and
place the measuring portion of the probe in air. Be
sure the probe is dry.
The adjacent screen will appear after selecting Zero
Cal from the Conductivity Calibration screen:
The adjacent screen will appear if zero Cal is successful.
The screen will return to the conductivity Cal Menu.
SECTION 7.0
CALIBRATION
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Cell Constant
1.00000 /cm
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Zero Cal
In Air
In Water
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Zero Cal
Sensor Zero Done
The adjacent screen may appear if zero Cal is unsuccessful.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Zero Cal
Sensor Zero Fail
Offset too high
Press EXIT
7.4.4 CALIBRATING THE SENSOR IN A CONDUCTIVITY
STANDARD (IN PROCESS CAL)
This procedure is used to calibrate the sensor and
analyzer against a solution of known conductivity.
This is done by submerging the probe in the sample of
known conductivity, then adjusting the displayed value, if
necessary, to correspond to the conductivity value of
the sample. Turn temperature correction off and use the
conductivity of the standard. Use a calibrated ther
mometer to measure temperature. The probe must be
cleaned before performing this procedure.
The adjacent screen will appear after selecting In
Process Cal from the Conductivity Calibration screen:
80
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN InProcess Cal
Wait for stable
reading.
MODEL 1056
SECTION 7.0
CALIBRATION
The adjacent screen will appear if In Process Cal is suc
cessful. The screen will return to the conductivity Cal
Menu.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
The adjacent screen may appear if In Process Cal is
unsuccessful. The screen will return to the conductivity
Cal Men
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN InProcess Cal
Updated cell
constant:
1.00135/cm
SN InProcess Cal
Calibration
Error
Press EXIT
7.4.5 CALIBRATING THE SENSOR TO A LABORATORY
INSTRUMENT (METER CAL)
This procedure is used to check and correct the
conductivity reading of the Model 1056 using a laboratory
conductivity instrument. This is done by submerging the
conductivity probe in a bath and measuring the conduc
tivity of a grab sample of the same bath water with a
separate laboratory instrument. The Model 1056 reading
is then adjusted to match the conductivity reading of the
lab instrument.
The adjacent screen will appear after selecting Meter
Cal from the Conductivity Calibration screen:
After pressing ENTER, the display shows the live value
measured by the sensor
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Meter Cal
Use precision
resistors only
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Enter Value
xx.xx kΩ
If the meter cal is successful the screen will return to the
conductivity Cal Menu.
The adjacent screen will appear if Meter Cal is unsuccessful.
The screen will return to the conductivity Cal Menu.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Meter Cal
Calibration
Error
Press EXIT
7.4.6 Cal Factor
Upon initial installation and power up, if 4electrode
was selected for the sensor type in the Quick Start
menus, the user enters a Cell Constant and a “Cal
Factor” using the instrument keypad. The cell constant
is needed to convert measured conductance to conduc
tivity as displayed on the analyzer screen. The “Cal
Factor” entry is needed increase the accuracy of the live
conductivity readings, especially at low conductivity
readings below 20uS/cm. Both the Cell Constant and
the “Cal Factor” are printed on the tag attached to the
4electrode sensor/cable.
The display screen for entering Cal Factor is shown.
The default value is displayed in bold type. If neces
sary after initial installation and startup, enter the “Cal
Factor” as printed on the sensor tag.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Cal Factor
0.95000 /cm
81
MODEL 1056
SECTION 7.0
CALIBRATION
7.5 TOROIDAL CONDUCTIVITY CALIBRATION
7.5.1 DESCRIPTION
Calibration is the process of adjusting or standardizing the analyzer to a lab test or a calibrated laboratory instru
ment, or standardizing to some known reference (such as a conductivity standard). This section contains procedures
for the first time use and for routine calibration of the Model 1056 analyzer.
THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1056 WITH AN ATTACHED INDUC
TIVE/TOROIDAL CONDUCTIVITY SENSOR. THE FOLLOWING CALIBRATION ROUTINES ARE COVERED
TABLE 74
Measure
Toroidal
Conductivity
Toroidal Conductivity Calibration
Sec.
Calibration function: default value
Description
7.5.2
Cell K:
3.00000/cm Enter the cell Constant for the sensor
7.5.3
Zero Cal
Zeroing the analyzer with the sensor attached
7.5.4
In Process Cal
Standardizing the sensor to a known conductivity
A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines.
To calibrate toroidal conductivity:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to
Toroidal Conductivity. Press ENTER.
4. Select Conductivity. Press ENTER.
The adjacent screen will appear. To calibrate Toroidal
Conductivity or Temperature, scroll to the desired item
and press ENTER
The following subsections show the initial display
screen that appears for each calibration routine. Use
the flow diagram for Conductivity calibration at the
end of Sec. 7 and the live screen prompts to complete
calibration.
The adjacent screen appears after selecting
Conductivity calibration:
82
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibrate?
Conductivity
Temperature
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibration
Zero Cal
In Process Cal
Cell K:
1.00000/cm
MODEL 1056
SECTION 7.0
CALIBRATION
7.5.2 ENTERING THE CELL CONSTANT
New conductivity sensors rarely need calibration. The
cell constant printed on the label is sufficiently accurate
for most applications. The cell constant should be
entered:
• When the unit is installed for the first time
• When the probe is replaced
• During troubleshooting
This procedure sets up the analyzer for the probe type
connected to the analyzer. Each type of probe has a
specific cell constant:
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Cell Constant
3.00000 /cm
The display screen for entering a cell constant for the
sensor is shown. The default value is displayed in bold
type.
7.5.3 ZEROING THE INSTRUMENT
This procedure is used to compensate for small offsets
to the conductivity signal that are present even when
there is no conductivity to be measured. This procedure
is affected by the length of extension cable and should
always be repeated if any changes in extension cable
or sensor have been made. Electrically connect the
conductivity probe as it will actually be used and place
the measuring portion of the probe in air.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Zero Cal
In Air
In Water
The adjacent screen will appear after selecting Zero
Cal from the Conductivity Calibration screen:
The adjacent screen will appear if zero Cal is successful.
The screen will return to the conductivity Cal Menu.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Zero Cal
Sensor Zero Done
.
The adjacent screen may appear if zero Cal is unsuccessful.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Zero Cal
Sensor Zero Fail
Offset too high
Press EXIT
7.5.4 CALIBRATING THE SENSOR IN A CONDUCTIVITY
STANDARD (IN PROCESS CAL)
This procedure is used to check and correct the
conductivity reading of the Model 1056 to ensure that
the reading is accurate. This is done by submerging the
probe in the sample of known conductivity, then adjusting
the displayed value, if necessary, to correspond to the
conductivity value of the sample. The probe must be
cleaned before performing this procedure. The temper
ature reading must also be checked and standardized if
necessary, prior to performing this procedure.
The adjacent screen will appear after selecting In
Process Cal from the Conductivity Calibration screen:
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN InProcess Cal
Wait for stable
reading.
83
MODEL 1056
The following screen will appear if In Process Cal is
successful. The screen will return to the conductivity
Cal Menu.
This screen may appear if In Process Cal is unsuccessful.
The screen will return to the conductivity Cal Menu.
SECTION 7.0
CALIBRATION
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN InProcess Cal
Updated cell
constant:
3.01350/cm
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN InProcess Cal
Calibration
Error
Press EXIT
7.6 CALIBRATION —CHLORINE
With a Chlorine measurement board and the appropriate sensor, Model 1056 can measure any of four variants of
Chlorine:
• Free Chlorine
• Total Chlorine
• Monochloramine
• pHindependent Free Chlorine
The section describes how to calibrate any compatible amperometric chlorine sensor. The following calibration
routines are covered in the family of supported Chlorine sensors:
• Air Cal
• Zero Cal
• In Process Cal
7.6.1 CALIBRATION — FREE CHLORINE
7.6.1.1 DESCRIPTION
A free chlorine sensor generates a current directly proportional to the concentration of free chlorine in the sam
ple. Calibrating the sensor requires exposing it to a solution containing no chlorine (zero standard) and to a solu
tion containing a known amount of chlorine (fullscale standard). The zero calibration is necessary because chlo
rine sensors, even when no chlorine is in the sample, generate a small current called the residual current. The
analyzer compensates for the residual current by subtracting it from the measured current before converting the
result to a chlorine value. New sensors require zeroing before being placed in service, and sensors should be
zeroed whenever the electrolyte solution is replaced. Either of the following makes a good zero standard:
• Deionized water containing about 500 ppm sodium chloride. Dissolve 0.5 grams (1/8 teaspoonful) of table
salt in 1 liter of water. DO NOT USE DEIONIZED WATER ALONE FOR ZEROING THE SENSOR. THE
CONDUCTIVITY OF THE ZERO WATER MUST BE GREATER THAN 50 μS/cm.
• Tap water known to contain no chlorine. Expose tap water to bright sunlight for at least 24 hours.
The purpose of the In Process calibration is to establish the slope of the calibration curve. Because stable chlorine
standards do not exist, the sensor must be calibrated against a test run on a grab sample of the process liquid.
Several manufacturers offer portable test kits for this purpose. Observe the following precautions when taking
and testing the grab sample.
• Take the grab sample from a point as close to the sensor as possible. Be sure that taking the sample does not alter
the flow of the sample to the sensor. It is best to install the sample tap just downstream from the sensor.
• Chlorine solutions are unstable. Run the test immediately after taking the sample. Try to calibrate the sensor
when the chlorine concentration is at the upper end of the normal operating range.
THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1056 WITH A FREE CHLORINE SENSOR.
THE FOLLOWING CALIBRATION ROUTINES ARE COVERED.
84
MODEL 1056
TABLE 75
Measure
Free Chlorine
SECTION 7.0
CALIBRATION
Free Chlorine Calibration Routines
Sec.
Calibration function: default value
Description
7.6.1.2
Zero Cal
Zeroing the sensor in solution with zero free chlorine
7.6.1.3
In Process Cal
Standardizing to a sample of known chlorine concentration
A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines.
To calibrate free chlorine:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to
Free Chlorine. Press ENTER.
4. Select Free Chlorine. Press ENTER.
The adjacent screen will appear. To calibrate Free
Chlorine or Temperature, scroll to the desired item and
press ENTER.
The following subsections show the initial display
screen that appears for each calibration routine. Use
the flow diagram for Chlorine calibration at the end
of Sec. 7 and the live screen prompts to complete cal
ibration.
The adjacent screen appears after selecting Free
Chlorine calibration:
7.6.1.2 ZEROING THE SENSOR.
The adjacent screen will appear during Zero Cal. Be
sure sensor has been running in zero solution for at
least two hours before starting zero step.
The adjacent screen will appear if In Zero Cal is
successful. The screen will return to the Amperometric
Cal Menu.
The adjacent screen may appear if In Zero Cal is unsuc
cessful. The screen will return to the Amperometric Cal
Menu.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibrate?
Free Chlorine
Temperature
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibration
Zero Cal
In Process Cal
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Zero Cal
Zeroing
Wait
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Zero Cal
Sensor zero done
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Zero Cal
Sensor zero failed
Press EXIT
7.6.1.3 IN PROCESS CALIBRATION
The adjacent screen will appear prior to In Process Cal
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN InProcess Cal
Wait for stable
reading.
85
MODEL 1056
SECTION 7.0
CALIBRATION
If the In Process Cal is successful, the screen will return
to the Cal submenu.
The adjacent screen may appear if In Zero Cal is unsuc
cessful. The screen will return to the Amperometric Cal
Menu.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN InProcess Cal
Calibration
Error
Press EXIT
7.6.2 CALIBRATION — TOTAL CHLORINE
7.6.2.1 DESCRIPTION
Total chlorine is the sum of free and combined chlorine.
The continuous determination of total chlorine requires
two steps. First, the sample flows into a conditioning
system (TCL) where a pump continuously adds acetic
acid and potassium iodide to the sample. The acid
lowers the pH, which allows total chlorine in the sample
to quantitatively oxidize the iodide in the reagent to
iodine. In the second step, the treated sample flows
to the sensor. The sensor is a membranecovered
amperometric sensor, whose output is proportional to
the concentration of iodine. Because the concentration
of iodine is proportional to the concentration of total
chlorine, the analyzer can be calibrated to read total
chlorine. Because the sensor really measures iodine,
calibrating the sensor requires exposing it to a solution
containing no iodine (zero standard) and to a solution
containing a known amount of iodine (fullscale standard).
The Zero calibration is necessary because the sensor,
even when no iodine is present, generates a small
current called the residual current. The analyzer
compensates for the residual current by subtracting it
from the measured current before converting the result
to a total chlorine value. New sensors require zeroing
before being placed in service, and sensors should be
TABLE 7 6
Measure
Total Chlorine
zeroed whenever the electrolyte solution is replaced.
The best zero standard is deionized water.
The purpose of the In Process Calibration is to
establish the slope of the calibration curve.
Because stable total chlorine standards do not exist, the
sensor must be calibrated against a test run on a
grab sample of the process liquid. Several manufac
turers offer portable test kits for this purpose. Observe
the following
precautions when taking and testing the grab sample:
• Take the grab sample from a point as close as
possible to the inlet of the TCL sample conditioning system.
Be sure that taking the sample does not alter the flow
through the TCL.
• Chlorine solutions are unstable. Run the test immedi
ately after taking the sample. Try to calibrate the sensor
when the chlorine concentration is at the upper end of
the normal operating range.
Note this measurement must be made using the Model
TCL total chlorine sample conditioning system.
THIS SECTION DESCRIBES HOW TO CALIBRATE
THE MODEL 1056 WITH AN ATTACHED TOTAL
CHLORINE SENSOR.
THE FOLLOWING
CALIBRATION ROUTINES ARE COVERED.
Total Chlorine Calibration Routines
Sec.
Calibration function: default value
Description
7.6.2.2
Zero Cal
Zeroing the sensor in solution with zero total chlorine
7.6.2.3
In Process Cal
Standardizing to a sample of known chlorine concentration
A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines.
To calibrate total chlorine:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to
Total Chlorine. Press ENTER.
4. Select Total Chlorine. Press ENTER.
The adjacent screen will appear. To calibrate Total
Chlorine or Temperature, scroll to the desired item and
press ENTER
86
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibrate?
Total Chlorine
Temperature
MODEL 1056
The following subsections provide you with the initial
display screen that appears for each calibration routine.
Use the flow diagram for Chlorine calibration at the
end of Sec. 7 and the live screen prompts to complete
calibration.
SECTION 7.0
CALIBRATION
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibration
Zero Cal
In Process Cal
This adjacent screen appears after selecting Total
Chlorine calibration:
7.6.2.2 ZEROING THE SENSOR.
The adjacent screen will appear during Zero Cal. Be
sure sensor has been running in zero solution for at
least two hours before starting zero step.
The adjacent screen will appear if In Zero Cal is suc
cessful. The screen will return to the Amperometric
Cal Menu.
The adjacent screen may appear if In Zero Cal is
unsuccessful.
The screen will return to the
Amperometric Cal Menu.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Zero Cal
Zeroing
Wait
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Zero Cal
Sensor zero done
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Zero Cal
Sensor zero failed
Press EXIT
7.6.2.3 IN PROCESS CALIBRATION
The adjacent screen will appear prior to In Process Cal
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN InProcess Cal
Wait for stable
reading.
If the In Process Cal is successful, the screen will
return to the Cal submenu.
The adjacent screen may appear if In Process Cal is
unsuccessful.
The screen will return to the
Amperometric Cal Menu.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN InProcess Cal
Calibration error
Press EXIT
87
MODEL 1056
SECTION 7.0
CALIBRATION
7.6.3 CALIBRATION MONOCHLORAMINE
7.6.3.1 DESCRIPTION
A monochloramine sensor generates a current directly
proportional to the concentration of monochloramine in
the sample. Calibrating the sensor requires exposing it
to a solution containing no monochloramine (zero
standard) and to a solution containing a known amount
of monochloramine (fullscale standard). The Zero
calibration is necessary because monochloramine
sensors, even when no monochloramine is in the
sample, generate a small current called the residual or
zero current. The analyzer compensates for the residual
current by subtracting it from the measured current
before converting the result to a monochloramine value.
New sensors require zeroing before being placed in
service, and sensors should be zeroed whenever the
electrolyte solution is replaced. The best zero standard
is deionized water.
The purpose of the In Process calibration is to establish
the slope of the calibration curve. Because stable
monochloramine standards do not exist, the sensor
must be calibrated against a test run on a grab sample
of the process liquid. Several manufacturers offer
portable test kits for this purpose. Observe the following
precautions when taking and testing the grab sample.
• Take the grab sample from a point as close to the sen
sor as possible. Be sure that taking the sample does not
alter the flow of the sample to the sensor. It is best to
install the sample tap just downstream from the sensor.
• Monochloramine solutions are moderately unstable.
Run the test as soon as possible after taking the sam
ple. Try to calibrate the sensor when the monochlo
ramine concentration is at the upper end of the normal
operating range.
THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1056 WITH AN ATTACHED MONO
CHLORAMINE SENSOR. THE FOLLOWING CALIBRATION ROUTINES ARE COVERED.
TABLE 77
Monochloramine Calibration Routines
Measure
Monochloramine
7.6.3.2
Zero Cal
Zeroing the sensor in solution with zero monochloramine
7.6.3.3
In Process Cal
Standardizing to a sample of known chlorine concentration
Sec.
Calibration function: default value
Description
A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines.
To calibrate monochloramine:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to
Monochloramine. Press ENTER.
4. Select Monochloramine. Press ENTER.
The adjacent screen will appear.
To calibrate
Monochloramine or Temperature, scroll to the desired
item and press ENTER.
The following subsections provide you with the initial
display screen that appears for each calibration routine.
Use the flow diagram for Chlorine calibration at the
end of Sec. 7 and the live screen prompts to complete
calibration.
The adjacent screen appears
Monochloramine calibration:
88
after
selecting
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibrate?
Monochloramine
Temperature
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibration
Zero Cal
In Process Cal
MODEL 1056
7.6.3.2 ZEROING THE SENSOR.
The adjacent screen will appear during Zero Cal. Be
sure sensor has been running in zero solution for at
least two hours before starting zero step.
The adjacent screen will appear if In Zero Cal is suc
cessful. The screen will return to the Amperometric
Cal Menu.
The adjacent screen may appear if In Zero Cal is
unsuccessful. The screen will return to the
Amperometric Cal Menu.
SECTION 7.0
CALIBRATION
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Zero Cal
Zeroing
Wait
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Zero Cal
Sensor zero done
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Zero Cal
Sensor zero failed
Press EXIT
7.6.3.3 IN PROCESS CALIBRATION
The adjacent screen will appear prior to In Process Cal
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN InProcess Cal
Wait for stable
reading.
If the In Process Cal is successful, the screen will
return to the Cal submenu.
The adjacent screen may appear if In Process Cal is
unsuccessful.
The screen will return to the
Amperometric Cal Menu.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN InProcess Cal
Calibration
Error
Press EXIT
89
MODEL 1056
SECTION 7.0
CALIBRATION
7.6.4 pHINDEPENDENT FREE CHLORINE
MEASUREMENT
7.6.4.1 DESCRIPTION
A free chlorine sensor generates a current directly
proportional to the concentration of free chlorine in the
sample. Calibrating the sensor requires exposing it to a
solution containing no chlorine (zero standard) and to a
solution containing a known amount of chlorine (full
scale standard). The zero calibration is necessary
because chlorine sensors, even when no chlorine is in
the sample, generate a small current called the residual
current. The analyzer compensates for the residual
current by subtracting it from the measured current
before converting the result to a chlorine value. New
sensors require zeroing before being placed in service,
and sensors should be zeroed whenever the electrolyte
solution is replaced. Either of the following makes a
good zero standard:
• Deionized water.
• Tap water known to contain no chlorine. Expose tap
water to bright sunlight for at least 24 hours.
The purpose of the In Process calibration is to
establish the slope of the calibration curve. Because
stable chlorine standards do not exist, the sensor must
be calibrated against a test run on a grab sample of
the process liquid.
Several manufacturers offer portable test kits for this
purpose. Observe the following precautions when taking
and testing the grab sample.
• Take the grab sample from a point as close to the sensor
as possible. Be sure that taking the sample does
not alter the flow of the sample to the sensor. It is best
to install the sample tap just downstream from the
sensor.
• Chlorine solutions are unstable. Run the test immediately
after taking the sample. Try to calibrate the sensor
when the chlorine concentration is at the upper end of
the normal operating range.
Note: This measurement is made using the model
498CL01 pHindependent Free Chlorine sensor
manufactured by Rosemount Analytical.
THIS SECTION DESCRIBES HOW TO CALIBRATE T H E M O D E L 1 0 5 6 W I T H A N AT TA C H E D PH
INDEPENDENT FREE CHLORINE SENSOR. THE FOLLOWING CALIBRATION ROUTINES ARE COVERED.
TABLE 7 8
Measure
pHindependent
Free Chlorine
pHindependent Free Chlorine Calibration Routines
Sec.
Calibration function: default value
Description
7.6.4.2
Zero Cal
Zeroing the sensor in solution with zero free chlorine
7.6.4.3
In Process Cal
Standardizing to a sample of known chlorine concentration
A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines.
To calibrate pHindependent free chlorine:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to
pHindependent free chlorine. Press ENTER.
4. Select pH Ind. Free Cl. Press ENTER.
The adjacent screen will appear. To calibrate pHindependent
Free Chlorine or Temperature, scroll to the desired item and
press ENTER.
90
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibrate?
pH Ind. Free Cl
Temperature
MODEL 1056
The following subsections provide you with the initial
display screen that appears for each calibration
routine. Use the flow diagram for Chlorine
calibration at the end of Sec. 7 and the live screen
prompts to complete calibration.
The adjacent screen appears after selecting pHinde
pendent free chlorine calibration:
7.6.4.2 ZEROING THE SENSOR.
The adjacent screen will appear during Zero Cal
SECTION 7.0
CALIBRATION
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibration
Zero Cal
In Process Cal
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Zero Cal
Zeroing
Wait
The adjacent screen will appear if In Zero Cal is suc
cessful. The screen will return to the Amperometric
Cal Menu.
The adjacent screen may appear if In Zero Cal is
unsuccessful.
The screen will return to the
Amperometric Cal Menu.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Zero Cal
Sensor zero done
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Zero Cal
Sensor zero failed
Press EXIT
7.6.4.3 IN PROCESS CALIBRATION
The following screen will appear prior to In Process Cal
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN InProcess Cal
Wait for stable
reading.
If the In Process Cal is successful, the screen will
return to the Cal submenu.
The adjacent screen may appear if In Process Cal is
unsuccessful.
The screen will return to the
Amperometric Cal Menu.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN InProcess Cal
Calibration
Error
Press EXIT
91
MODEL 1056
7.7 CALIBRATION — OXYGEN
7.7.1 DESCRIPTION
Oxygen sensors generate a current directly proportional
to the concentration of dissolved oxygen in the sample.
Calibrating the sensor requires exposing it to a solution
containing no oxygen (zero standard) and to a solution
containing a known amount of oxygen (fullscale
standard). The Zero Calibration is necessary because
oxygen sensors, even when no oxygen is present in the
sample, generate a small current called the residual
current. The analyzer compensates for the residual
current by subtracting it from the measured current
before converting the result to a dissolved oxygen
value. New sensors require zeroing before being placed
in service, and sensors should be zeroed whenever the
electrolyte solution is replaced. The recommended zero
standard is 5% sodium sulfite in water, although
oxygenfree nitrogen can also be used. The Model
499A TrDO sensor, used for the determination of
trace (ppb) oxygen levels, has very low residual
current and does not normally require zeroing. The
residual current in the 499A TrDO sensor is equivalent
to less than 0.5 ppb oxygen.
The purpose of the In Process Calibration is to
establish the slope of the calibration curve. Because the
solubility of atmospheric oxygen in water as a function
of temperature and barometric pressure is well known,
the natural choice for a fullscale standard is airsaturated
water. However, airsaturated water is difficult to
prepare and use, so the universal practice is to use air
for calibration. From the point of view of the oxygen
sensor, air and airsaturated water are identical. The
equivalence comes about because the sensor really
measures the chemical potential of oxygen. Chemical
potential is the force that causes oxygen molecules to
diffuse from the sample into the sensor where they can
be measured. It is also the force that causes oxygen
molecules in air to dissolve in water and to continue to
dissolve until the water is saturated with oxygen. Once
the water is saturated, the chemical potential of oxygen
in the two phases (air and water) is the same.
Oxygen sensors generate a current directly proportional
to the rate at which oxygen molecules diffuse through a
membrane stretched over the end of the sensor. The
92
SECTION 7.0
CALIBRATION
diffusion rate depends on the difference in chemical
potential between oxygen in the sensor and oxygen in
the sample. An electrochemical reaction, which
destroys any oxygen molecules entering the sensor,
keeps the concentration (and the chemical potential) of
oxygen inside the sensor equal to zero. Therefore, the
chemical potential of oxygen in the sample alone deter
mines the diffusion rate and the sensor current.
When the sensor is calibrated, the chemical potential of
oxygen in the standard determines the sensor current.
Whether the sensor is calibrated in air or airsaturated
water is immaterial. The chemical potential of oxygen
is the same in either phase. Normally, to make the
calculation of solubility in common units (like ppm DO)
simpler, it is convenient to use watersaturated air for
calibration. Automatic air calibration is standard. The
user simply exposes the sensor to watersaturated air.
The analyzer monitors the sensor current. When the
current is stable, the analyzer stores the current and
measures the temperature using a temperature element
inside the oxygen sensor. The user must enter the
barometric pressure.
From the temperature the analyzer calculates the
saturation vapor pressure of water. Next, it calculates
the pressure of dry air by subtracting the vapor pressure
from the barometric pressure. Using the fact that dry air
always contains 20.95% oxygen, the analyzer calcu
lates the partial pressure of oxygen. Once the analyzer
knows the partial pressure of oxygen, it uses the
Bunsen coefficient to calculate the equilibrium solubility
of atmospheric oxygen in water at the prevailing tem
perature. At 25°C and 760 mm Hg, the equilibrium sol
ubility is 8.24 ppm. Often it is too difficult or messy to
remove the sensor from the process liquid for calibra
tion. In this case, the sensor can be calibrated against
a measurement made with a portable laboratory instru
ment. The laboratory instrument typically uses a mem
branecovered amperometric sensor that has been cali
brated against watersaturated air.
THIS SECTION DESCRIBES HOW TO CALIBRATE
THE MODEL 1056 WITH AN OXYGEN SENSOR.
THE FOLLOWING CALIBRATION ROUTINES ARE
COVERED.
MODEL 1056
TABLE 7 9
Measure
Oxygen
SECTION 7.0
CALIBRATION
Oxygen Calibration Routines
Sec.
Calibration function: default value
Description
7.7.2
Zero Cal
Zeroing the sensor in a medium with zero oxygen
7.7.3
Air Cal
Calibrating the sensor in a watersaturated air sample
7.7.4
In Process Cal
Standardizing to a sample of known oxygen concentration
7.7.5
Sen@ 25°C:2500nA/ppm
Entering a known slope value for sensor response
7.7.6
Zero Current:
0nA Entering a known zero current for a specific sensor
A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines.
Oxygen sensors generate a current directly proportional
to the concentration of dissolved oxygen in the sample.
Calibrating the sensor requires exposing it to a solution
containing no oxygen (zero standard) and to a solution
containing a known amount of oxygen (fullscale standard).
Automatic air calibration is standard. The user simply
exposes the sensor to watersaturated air.
To calibrate oxygen:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to
oxygen. Press ENTER.
4. Select Oxygen. Press ENTER.
The adjacent screen will appear. To calibrate Oxygen
or Temperature, scroll to the desired item and press
ENTER
The following subsections provide you with the initial
display screen that appears for each calibration routine.
Use the flow diagram for Oxygen calibration at the
end of Sec. 7 and the live screen prompts for each routine
to complete calibration.
The adjacent screen appears after selecting Oxygen
calibration:
Air calibration criteria can be changed. The following
criteria can be adjusted:
Stabilization time (default 10 sec.)
Stabilization pH value (default 0.05 ppm)
Salinity of the solution to be measured (default
00.0 parts per thousand)
The adjacent screen will appear to allow adjustment of
these criteria”
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibrate?
Oxygen
Temperature
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibration
Air Cal
Zero Cal
In Process Cal
Sen@ 25°C:2500nA/ppm
Zero Current: 1234nA
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Setup
Stable Time: 10 sec
Stable Delta: 0.05 ppm
Salinity: 00.0 ‰
93
MODEL 1056
SECTION 7.0
CALIBRATION
7.7.2 ZEROING THE SENSOR.
The adjacent screen will appear during Zero Cal
.
S1:
1.234 nA
S2:
1.456 nA
SN Zero Cal
Zeroing
Wait
The adjacent screen will appear if In Zero Cal is suc
cessful. The screen will return to the Amperometric
Cal Menu.
S1:
1.234 nA
S2:
1.456 nA
The adjacent screen may appear if In Zero Cal is
unsuccessful.
The screen will return to the
Amperometric Cal Menu.
S1:
1.234 nA
S2:
1.456 nA
SN Zero Cal
Sensor zero done
SN Zero Cal
Sensor zero failed
Press EXIT
7.7.3 CALIBRATING THE SENSOR IN AIR
The adjacent screen will appear prior to Air Cal
S1: 1.234µS/cm
S2: 12.34pH
25.0ºC
25.0ºC
SN Air Cal
Start Calibration
Setup
The adjacent screen will appear if In Air Cal is suc
cessful. The screen will return to the Amperometric
Cal Menu.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
The adjacent screen will appear if In Air Cal is suc
cessful. The screen will return to the Amperometric
Cal Menu.
S1: 1.234µS/cm
S2: 12.34pH
SN Air Cal
Done
25.0ºC
25.0ºC
SN Air Cal
Failure
Check Sensor
Press EXIT
7.7.4 CALIBRATING THE SENSOR AGAINST A
STANDARD INSTRUMENT (IN PROCESS CAL)
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
The adjacent screen will appear prior to In Process Cal
SN InProcess Cal
Wait for stable
reading.
If the In Process Cal is successful, the screen will
return to the Cal submenu.
S1: 1.234µS/cm
S2: 12.34pH
The adjacent screen may appear if In Zero Cal is
unsuccessful.
The screen will return to the
Amperometric Cal Menu.
94
25.0ºC
25.0ºC
SN InProcess Cal
Calibration
Error
Press EXIT
MODEL 1056
SECTION 7.0
CALIBRATION
7.8 CALIBRATION — OZONE
7.8.1 DESCRIPTION
An ozone sensor generates a current directly
proportional to the concentration of ozone in the sample.
Calibrating the sensor requires exposing it to a solution
containing no ozone (zero standard) and to a solution
containing a known amount of ozone (fullscale
standard). The Zero Calibration is necessary because
ozone sensors, even when no ozone is in the sample,
generate a small current called the residual or zero
current. The analyzer compensates for the residual
current by subtracting it from the measured current
before converting the result to an ozone value. New
sensors require zeroing before being placed in service,
and sensors should be zeroed whenever the electrolyte
solution is replaced. The best zero standard is deionized
water.
against a test run on a grab sample of the process liquid.
Several manufacturers offer portable test kits for this
purpose. Observe the following precautions when taking
and testing the grab sample.
• Take the grab sample from a point as close to the sensor
as possible. Be sure that taking the sample does not
alter the flow of the sample to the sensor. It is best to
install the sample tap just downstream from the sensor.
• Ozone solutions are unstable. Run the test immedi
ately after taking the sample. Try to calibrate the sensor
when the ozone concentration is at the upper end of the
normal operating range.
THIS SECTION DESCRIBES HOW TO CALIBRATE
THE MODEL 1056 WITH AN OZONE SENSOR. THE
FOLLOWING CALIBRATION ROUTINES ARE COV
ERED.
The purpose of the In Process Calibration is to establish
the slope of the calibration curve. Because stable ozone
standards do not exist, the sensor must be calibrated
TABLE 7 10
Measure Ozone
Ozone Calibration Routines
Sec.
Calibration function: default value
Description
7.8.2
Zero Cal
Zeroing the sensor in solution with zero ozone
7.8.3
In Process Cal
Standardizing to a sample of known ozone concentration
A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines.
To calibrate ozone:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to
ozone. Press ENTER.
4. Select Ozone. Press ENTER.
The adjacent screen will appear. To calibrate Ozone or
Temperature, scroll to the desired item and press
ENTER.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibrate?
Ozone
Temperature
95
MODEL 1056
The following subsections provide you with the
initial display screen that appears for each calibration
routine. Use the flow diagram for Ozone calibration
at the end of Sec. 7 and the live screen prompts to
complete calibration.
SECTION 7.0
CALIBRATION
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibration
Zero Cal
In Process Cal
The adjacent screen appears after selecting Ozone
calibration:
7.8.2 ZEROING THE SENSOR.
The following screen will appear during Zero Cal
S1:
S2:
The following screen will appear if In Zero Cal is
successful.
The screen will return to the
Amperometric Cal Menu.
S1:
1.234 nA
S2:
1.456 nA
The following screen may appear if In Zero Cal is
unsuccessful.
The screen will return to the
Amperometric Cal Menu.
S1:
S2:
1.234 nA
1.456 nA
SN Zero Cal
Zeroing
Wait
SN Zero Cal
Sensor zero done
1.234 nA
1.456 nA
SN Zero Cal
Sensor zero failed
Press EXIT
7.8.3 IN PROCESS CALIBRATION
The following screen will appear after selecting In
Process Cal
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
If the In Process Cal is successful, the screen will
return to the Cal submenu.
S1: 1.234µS/cm
S2: 12.34pH
The following screen may appear if In Zero Cal is
unsuccessful.
The screen will return to the
Amperometric Cal Menu.
96
SN InProcess Cal
Wait for stable
reading.
25.0ºC
25.0ºC
SN InProcess Cal
Calibration
Error
Press EXIT
MODEL 1056
SECTION 7.0
CALIBRATION
7.9 CALIBRATING TEMPERATURE
7.9.1 DESCRIPTION
Most liquid analytical measurements require temperature compensation (except ORP). The Model 1056 performs
temperature compensation automatically by applying internal temperature correction algorithms. Temperature cor
rection can also be turned off. If temperature correction is off, the Model 1056 uses the manual temperature entered by
the user in all temperature correction calculations.
THIS SECTION DESCRIBES HOW TO CALIBRATE TEMPERATURE IN THE MODEL 1056 ANALYZER. THE
FOLLOWING CALIBRATION ROUTINE IS COVERED.
TABLE 7 11
Measure
Temperature
Temperature Calibration Routine
Sec.
7.9.2
Calibration function: default value
Calibrate
Description
Enter a manual reference temperature for temperature
compensation of the process measurement
A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines.
To calibrate temperature:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to
the desired measurement. Press ENTER.
4. Select Temperature. Press ENTER.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibrate
+025.0°C
The adjacent screen will appear.
The following subsection provides you with the initial
display screen that appears for temperature calibration.
Use the flow diagram for Temp calibration at the
end of Sec. 7 to complete calibration.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibrate
Cal in progress.
Please wait.
7.9.2 CALIBRATION
The adjacent screen will appear during Temperature
Cal.
If the sensor Temperature offset is greater than 5 ºC
from the default value, the following screen will
appear:
S1: 1.234µS/cm
S2: 12.34pH
You may continue by selecting Yes or suspend this
operation by selecting No.
No
Yes
25.0ºC
25.0ºC
SN Temp Offset > 5°C
Continue?
If the Temp Cal is successful, the screen will return to
the Cal Menu.
Note: To select automatic or manual temp compensation
or to program temperature units as °C or °F, refer to
Sec. 5.3 – Programming Temperature in this manual
97
MODEL 1056
SECTION 7.0
CALIBRATION
7.10 TURBIDITY
7.10.1 DESCRIPTION
This section describes how to calibrate the turbidity sensor against a userprepared standard as a 2point calibra
tion with diionized water, against a 20 NTU userprepared standard as a single point calibration, and against a
grab sample using a reference turbidimeter.
THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1056 WITH AN ATTACHED TURBIDITY
SENSOR AS PART OF THE COMPLETE CLARITY II TURBIDITY SYSTEM. THE FOLLOWING CALIBRATION
ROUTINES ARE COVERED.
TABLE 712 TURBIDITY CALIBRATION ROUTINES
Measure Sec. Calibration function:
Turbidity 7.10.2 Slope Calibration
default value
Description
Slope cal with pure water and a standard of known turbidity
7.10.3 Standardize Calibration
Standardizing the sensor to a known turbidity
7.10.4 Grab Calibration
Standardizing the sensor to a known turbidity based on a
reference turbidimeter
A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines.
To calibrate Turbidity:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to
Turbidity. Press ENTER.
4. Select Turbidity. Press ENTER.
The following subsections provide you with the initial
display screen that appears for each calibration rou
tine. Use the flow diagram for Turbidity calibration
at the end of Sec. 7 and the live screen prompts to
complete calibration.
7.10.2 SLOPE CALIBRATION — Turbidity
This section describes how to conduct a 2point cali
bration of the turbidity sensor against a userprepared
20NTU standard. The calibration requires two steps.
First, immerse the sensor in filtered water having very
low turbidity and measure the sensor output. Next,
increase the turbidity of the filtered water by a known
amount, typically 20 NTU, and measure the sensor
output again. The analyzer takes the two measure
ments, applies a linearization correction (if necessary),
and calculates the sensitivity. Sensitivity is the sensor
output (in mV) divided by turbidity. A typical new sen
sor has a sensitivity of about 10 mV/NTU. As the sen
sor ages, the sensitivity decreases. The figure below
illustrates how turbidity calibration works. Before
beginning the calibration, the analyzer does a dark
current measurement. Dark current is the signal gener
ated by the detector when no light is falling on it. The
analyzer subtracts the dark current from the raw scat
98
The following screen will appear.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibrate?
Turbidity
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibrate
Slope
Standard
Grab
tered light signal and converts the result to turbidity. In
highly filtered samples, which scatter little light, the
dark current can be a substantial amount of the signal
generated by the detector.
MODEL 1056
This screen appears after selecting Slope calibration.
SECTION 7.0
CALIBRATION
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Slope Cal
Sensor in pure H2O?
Press ENTER
The following screen will appear if Slope Cal is suc
cessful. The screen will return to the Turbidity Cal
Menu.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
The following screen may appear if Slope Cal is
unsuccessful.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Slope Cal
Cal Complete
SN Slope Cal
Calibration
Error
Press EXIT
7.10.3 STANDARDIZE CALIBRATION Turbidity
The turbidity sensor can also be calibrated against a
commercial standard. Stable 20.0 NTU standards are
available from a number of sources. Calibration using a
commercial standard is simple. Filtered deionized water
is not required. Before beginning the calibration, the
analyzer does a dark current measurement. Dark cur
rent is the signal generated by the detector even when
no light is falling on it. The analyzer subtracts the dark
current from the raw scattered light signal and converts
the result to turbidity. In highly filtered samples, which
scatter little light, the dark current can be a substantial
amount of the signal generated by the sensor.
This screen appears after selecting Standard calibration.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Standard Cal
Sensor in Standard?
Press ENTER
The following screen will appear if Standard Cal is
successful. The screen will return to the Turbidity Cal
Menu.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
The following screen may appear if Standard Cal is
unsuccessful.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Standard Cal
Cal Complete
SN Standard Cal
Calibration
Error
Press EXIT
99
MODEL 1056
SECTION 7.0
CALIBRATION
7.10.4 GRAB CALIBRATION Turbidity
If desired, the turbidity sensor can be calibrated
against the turbidity reading from another instrument.
The analyzer treats the value entered by the user as
though it were the true turbidity of the sample.
Therefore, grab sample calibration changes the sensi
tivity, it does not apply an offset to the reading.
This screen appears after selecting Grab calibration.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Grab Cal
Wait for stable
reading
The following screen will appear if Grab Cal is successful.
The screen will return to the Turbidity Cal Menu.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Grab Cal
Cal Complete
The following screen may appear if Grab Cal is
unsuccessful.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Grab Cal
Calibration
Error
Press EXIT
7.11 PULSE FLOW
7.11.1 DESCRIPTION
A variety of pulse flow sensors can be wired to the Flow signal input board to measure flow volume, total volume
and flow difference (if 2 Flow signal boards are installed). The Model 1056 Flow signal board will support flow
sensors that are selfdriven (powered by the rotation of the impeller paddlewheel).
THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1056 WITH AN ATTACHED FLOW SENSOR.
THE FOLLOWING CALIBRATION ROUTINES ARE COVERED.
TABLE 713 FLOW CALIBRATION ROUTINES
Measure
Pulse
Flow
Calibration function
Description
7.11.2
K Factor
A constant value representing pulses/Gal of flow
7.11.3
Frequency/Velocity & Pipe
Alternate cal method – requires manual entry of frequency (Hz)
per velocity and Pipe diameter used
7.11.4
In process Calibration
Calibration based on known volume per unit of time
7.11.5
Totalizer Control
User settings to stop, restart and reset total volume meter
Sec.
A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines.
To
1.
2.
3.
calibrate Pulse Flow:
Press the MENU button
Select Calibrate. Press ENTER.
Select Sensor 1 or Sensor 2 corresponding to
Flow. Press ENTER.
4. Select Pulse Flow. Press ENTER.
The following screen will appear.
100
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibrate?
Pulse Flow
MODEL 1056
To calibrate Pulse Flow scroll to the desired item and
press ENTER.
The following subsections provide you with the initial
display screen that appears for each calibration rou
tine. Use the diagram for Pulse Flow calibration at
the end of Sec. 7 and the live screen prompts to com
plete calibration.
7.11.2 CALIBRATION — K Factor
This screen appears after selecting K Factor.
SECTION 7.0
CALIBRATION
.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibration
K Factor: 12.34 p/Gal
Freq/Velocity & Pipe
In Process
Totalizer Control
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN K Factor
12.34 p/Gal
.
To calibrate Pulse Flow scroll to the desired item and
press ENTER.
The following subsections provide you with the initial
display screen that appears for each calibration rou
tine. Use the diagram for Pulse Flow calibration at
the end of Sec. 7 and the live screen prompts to com
plete calibration.
.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibration
K Factor: 12.34 p/Gal
Freq/Velocity & Pipe
In Process
Totalizer Control
Simply enter the know K factor provided with the flow
sensor specifications. The screen will return to the
Pulse Flow Cal Menu and the updated K factor will
appear.
7.11.3 CALIBRATION — Freq/Velocity & Pipe
This screen appears after selecting Freq/Velocity &
Pipe calibration.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
After completing the entry of the Freq/Velocity ratio, the
following screen will appear:
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
The following screen will appear if the entries are suc
cessful. The screen will return to the Pulse Flow Cal
Menu.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Freq/Velocity
12.34 Hz per ft/sec
SN Pipe Diameter
10.00 in
SN Freq/Velocity&Pipe
Updated K Factor
12.34 p/Gal
101
MODEL 1056
The following screen may appear if the entries are
unsuccessful.
SECTION 7.0
CALIBRATION
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Freq/Velocity&Pipe
Calibration
Error
Press EXIT
7.11.4 CALIBRATION — In Process Cal
This screen appears after selecting In Process Cal.
The following screen will appear if the entries are suc
cessful. The screen will return to the Pulse Flow Cal
Menu.
The following screen will appear if the entries are suc
cessful.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN InProcess Cal
Updated K Factor
12.34 p/Gal
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN InProcess Cal
Calibration
Error
Press EXIT
7.11.5 CALIBRATION — Totalizer Control
This screen appears after selecting Totalizer Control.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Totalizer Control
Stop
Resume
Reset
123456789012.3 G
The user can suspend the totalizer by selecting Stop, reenable the totalizer by selecting Resume and return the
totalizer volume count to zero by selecting Reset. The live totalizer volume count is displayed during these
menu operations.
102
SECTION 7.0
CALIBRATION
FIGURE 71 Calibrate pH
MODEL 1056
103
FIGURE 72 Calibrate ORP
MODEL 1056
104
SECTION 7.0
CALIBRATION
FIGURE 73 Calibrate Contacting and Toroidal Conductivity
MODEL 1056
SECTION 7.0
CALIBRATION
105
FIGURE 74 Calibrate Free Chlorine, Total Chlorine, Monochloramine, and
pHindependent Free Chlorine
MODEL 1056
106
SECTION 7.0
CALIBRATION
SECTION 7.0
CALIBRATION
FIGURE 75 Calibrate Oxygen
MODEL 1056
107
FIGURE 76 Calibrate Ozone
MODEL 1056
108
SECTION 7.0
CALIBRATION
SECTION 7.0
CALIBRATION
FIGURE 77 Calibrate Temperature
MODEL 1056
109
FIGURE 78 Calibrate Turbidity
MODEL 1056
110
SECTION 7.0
CALIBRATION
SECTION 7.0
CALIBRATION
FIGURE 79 Calibrate Flow
MODEL 1056
111
MODEL 1056
SECTION 8.0
RETURN OF MATERIAL
SECTION 8.0
RETURN OF MATERIAL
17.1 GENERAL.
To expedite the repair and return of instruments, proper communication between the customer and the factory is
important. Before returning a product for repair, call 19497578500 for a Return Materials Authorization (RMA)
number.
17.2 WARRANTY REPAIR.
The following is the procedure for returning instruments still under warranty:
1.
Call Rosemount Analytical for authorization.
2.
To verify warranty, supply the factory sales order number or the original purchase order number. In the case
of individual parts or subassemblies, the serial number on the unit must be supplied.
3.
Carefully package the materials and enclose your “Letter of Transmittal” (see Warranty). If possible, pack the
materials in the same manner as they were received.
4.
Send the package prepaid to:
Emerson Process Management
Rosemount Analytical
2400 Barranca Parkway
Irvine, CA 92606
Attn: Factory Repair
RMA No. ____________
IMPORTANT
Please see second section of “Return of
Materials Request” form. Compliance with
the OSHA requirements is mandatory for
the safety of all personnel. MSDS forms
and a certification that the instruments have
been disinfected or detoxified are required.
Mark the package: Returned for Repair
Model No. ____
17.3 NONWARRANTY REPAIR.
The following is the procedure for returning for repair instruments that are no longer under warranty:
1.
Call Rosemount Analytical for authorization.
2.
Supply the purchase order number, and make sure to provide the name and telephone number of the individ
ual to be contacted should additional information be needed.
3.
Do Steps 3 and 4 of Section 17.2.
NOTE
Consult the factory for additional information regarding service or repair.
112
The right people, the right answers, right now.
Immediate, Reliable Analytical Support
Now there’s a way to quickly get the right answers for your liquid analytical instrumentation questions: the Analytical
Customer Support Center.
Our staff of trained professionals is ready to provide the information you need. If you are placing an order, verifying delivery,
requesting application information, or just want to contact a Rosemount Analytical representative, a call to the Customer
Support Center will provide you with the right people, the right answers, right now.
A Worldwide Network of Sales and Service
Emerson Process Management’s field sales offices are your source for more information on the fill line of Rosemount Analytical
products. Field sales personnel will work closely with you to supply technical data and application information.
For more information, please contact your nearest Emerson Process Management sales office.
THE AMERICAS
HEADQUARTERS
Emerson Process Management
Rosemount Analytical Inc.
Liquid Center of Excellence
2400 Barranca Parkway
Irvine, CA 92606
Phone: +1.949.757.8500
Toll Free: +1.800.854.8257
Fax: +1.949.474.7250
ASIAPACIFIC
Emerson Process Management
Asia Pacific Private Ltd.
1 Pandan Crescent
Singapore 0512
Republic of Singapore
Phone: 65.777.8211
Fax: 65.777.0947
GERMANY
Emerson Process Management Process
Gas Analyzer Center of Excellence
GmbH & Co. OHG
Industriestrasse 1
63594 Hasselroth
Germany
T 49.6055.884.0
F 49.6055.884.20
LATIN AMERICA
Emerson Process Management
Rosemount Analytical
10241 West Little York, Suite #200
Houston, TX 77040 USA
T 713.467.6000
F 713.827.3328
MIDDLE EAST AND AFRICA
EUROPE
Emerson Process Management AG
Blegistrasse 21
CH6341 BaarWalterswil
Switzerland
T 41.41.768.6111
T 41.41.761.8740
VISIT OUR WEBSITE AT
www.rosemountanalytical.com
Emerson Process Management
EPM Building
P. O. Box 17033
Jebe Ali Free Zone
Dubai, United Arab Emirates
T 971.4.8835235
F 971.4.8835312
The right people,
the right answers,
right now.
ONLINE ORDERING NOW AVAILABLE ON OUR WEB SITE
http://www.rosemountanalytical.com
8
Credit Cards for U.S. Purchases Only.
Emerson Process Management
2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 7578500
Fax: (949) 4747250
http://www.rosemountanalytical.com
© Rosemount Analytical Inc. 2014