HACH - WHAT IS pH AND HOW IS IT MEASURED? A Technical Handbook for Industry
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
      WHAT IS pH  AND HOW IS IT MEASURED?            A Technical Handbook for Industry                                                            CHAPTER 1 – INTRODUCTION  Why Is pH Measurement Necessary?  3  CHAPTER 2 – WATER AND AQUEOUS SOLUTIONS  The Properties of Water  Ion Product Constant of Water  Molarity  4    4  5  CHAPTER  3  –  FUNDAMENTALS  OF  pH  Definition of pH  pH Values and Hydrogen/Hydroxide Concentration  How is pH Measured?  Activity versus Concentration  The Nernst Equation  The Standard Hydrogen Electrode  6    6    7    7    9    9  CHAPTER 4 – THE pH SENSOR  pH Electrodes  The Measuring Electrode  Asymmetry Potential  Sodium Ion Error  Acid Error  Temperature  Effects  The Reference Electrode  The Reference Junction  Junction Potentials  Gel‐filled Reference Electrode  Buffers  Calibration  11    13  13  14    15  15   16    17  17    18    18  18  CHAPTER  5  –  CARE  OF  pH  ELECTRODES  Dehydration  Factors Detrimental to Electrode Life  Transportation  Storage  22    22  22   23  CHAPTER 6 – COMMON APPLICATIONS  Cooling Tower Control  Food Processing  Coal Industry  Plating Waste Treatment  Ultrapure Water    24    24           24          25          25  BIBLIOGRAPHY                                                                                                  27    GLOSSARY                                                                                             28    Chapter 1 – INTRODUCTION Why Is pH Measurement Necessary? Almost all processes containing water have a need for pH measurement. Most living things depend on a proper pH  level to sustain life. All human beings and animals rely on internal mechanisms to maintain the pH level of their  blood. The blood flowing through our veins must have a pH between 7.35 and 7.45. Exceeding this range by as little  as one‐tenth of a pH unit could prove fatal.    Commodities such as wheat and corn, along with other plants and food products, will grow best if the soil they are  planted in is maintained at an optimal pH. To attain high crop yields, farmers must condition their fields to the  correct pH value. Different crops need different pH levels. In this case, one size does not fit all.    Acid rain can be very detrimental to crop yields. Rainwater is naturally acidic (below 7.0 pH). Rain is typically  around 5.6 pH but, in some areas, it increases to harmful levels between 4.0 and 5.0 pH due to atmospheric  pollutants. Heavily industrialized areas of the US, such as the Midwest, have been targeted by various  environmental agencies to minimize the pollutants that cause acid rain. The burning of fossil fuels like coal,  releases gases into the upper atmosphere that, when combined with rain water, change composition and cause  the rain water to become more acidic.    Proper pH control keeps milk from turning sour, makes strawberry jelly gel, and prevents shampoo from stinging  your eyes. In plating plants, pH control is used to ensure the luster of chrome on various products from nuts and  bolts to toasters and automobile bumpers. The pH of wastewater leaving manufacturing plants and wastewater  purification plants, as well as potable water from municipal drinking water plants, must be within a specific pH  "window" as set forth by local, state or federal regulatory agencies. This value is typically between 5 and 9 pH, but  can vary from area to area.    Other pH applications include:  • Neutralization of effluent in steel, pulp and paper, chemical, and pharmaceutical manufacturing  • Hexavalent chromium destruction  • Cyanide destruction  • Reverse osmosis  • Odor scrubbers  • Pharmaceutical  manufacturing  • Chemical and petrochemical manufacturing  • Cooling tower control    Whether adjusting the pH for a proper reaction or making sure wastewater is at the proper pH value before  sending it to the community sewer system, accurate pH measurement is required. Put simply, pH is an integral part  of our everyday life.  Chapter 2 – WATER AND AQUEOUS SOLUTIONS   The Properties of Water Water is the most common substance known to man, and is the most important. In vapor, liquid or solid form,  water covers more than seventy percent of the Earth’s surface, and is a major component of the atmosphere.  Water is also an essential requirement for all forms of life. Most living things are largely made up of water. Human  beings, for example, consist of about two‐thirds water.    Pure water is a clear, colorless, and odorless liquid made  up of one oxygen and two hydrogen  atoms. The Italian  scientist Stanislao Cannizzarro defined the chemical formula of the water molecule, H2O, in 1860. Water is a very  powerful substance that acts as a medium for many reactions, which is why it is often referred to as the "universal  solvent." Although pure water is a poor conductor of electricity, impurities that occur naturally in water transform  it into a relatively good conductor. Water has unusually high boiling (100°C/212°F) and freezing (0°C/32°F) points. It  also shows unusual volume changes with temperature. As water cools, it contracts to a maximum density of 1 gram  per cubic centimeter at 4°C (39°F). Further cooling actually causes it to expand, especially when it reaches the  freezing point. The fact that water is denser in the liquid form than the solid form explains why an ice cube floats in  a beverage, or why a body of water freezes from the top down. While the density property of water is of little  importance to the beverage example, it has a tremendous impact on the survival of aquatic life inhabiting a body  of water.    Ion Product Constant of Water   Water  molecules  are  in  continuous  motion,  even  at  lower  temperatures.  When  two  water  molecules  collide,  a  hydrogen ion is transferred from one molecule to the other (Figure 1). The water molecule that loses the hydrogen  ion  becomes  a  negatively  charged  hydroxide  ion.  The  water  molecule  that  gains  the  hydrogen  ion  becomes  a  positively charged hydronium ion. This process is commonly referred to as the self‐ionization of water.    The self‐ionization of water does not regularly occur. This reaction can be written as a simple dissociation  (Figure 2).  At 25⁰ C in pure water, each concentration of hydrogen ions and hydroxide ions is only 1 x 10‐7 M. It  is important to note that the number of hydrogen and hydroxide ions produced from this reaction is equal. This  is why pure water is often described as a neutral solution.  In all other aqueous solutions, the relative concentrations of each of these ions is unequal. When more of one  ion is added to the solution, the concentration of the other decreases. The following equation describes this  relationship:        The product of the hydrogen and hydroxide ions is always equal to 1 x 10‐14 (mol/L)2. Therefore, if the  concentration of one ion increases by a factor of 10, then the concentration of the other ion must decrease by a  factor of 10. Since this relationship is constant, it is given the symbol KW, which is called the ion‐product  constant for water.    Aqueous solutions that have a hydrogen ion concentration greater than the hydroxide ion concentration are  called acidic solutions. When the hydroxide ion concentration is greater than the hydrogen ion concentration,  the solution is called basic or alkaline.    Molarity    The term "molarity" describes the concentration of a substance within a solution. By definition, a one "molar"  solution of hydrogen ion contains one "mole" of hydrogen ion per liter of solution. Therefore, a solution of 10 pH  has 1 x 10‐10 moles of hydrogen ions as shown by the following equation:           Furthermore, a solution of 4 pH has 1 x 10‐4 moles of hydrogen ions, and so on. This also means that one liter of  a pH 10 solution would contain 1x10‐10 grams of hydrogen ion, because 1 mole = 1 g/L for hydrogen.    A one molar solution of sodium hydroxide (NaOH), a base, is approximately 4% by weight, and has a pH value of  14. A one molar solution of hydrochloric acid (HCl), an acid, is approximately 3.7% by weight, and has a pH of 0.  By diluting either of these two solutions, the molarity will decrease as well. For example, diluting 1 ml of HCl  acid by adding 9 ml of distilled water results in a 0.1 molar hydrochloric acid solution, which has a pH value of  1.0. Diluting sodium hydroxide using the same volumes yields a solution with a pH value of 13. If this dilution  procedure were continued, the pH of each solution would approach a neutral pH of 7.     NOTE: For every 10‐fold change in concentration (example: 0.1 to 1.0), the pH changes by one unit.  If equal volumes of 4 pH (0.0001M HCl) and 10 pH (0.0001 NaOH) solutions were mixed together, the resultant  solution would have a pH of 7.    NOTE: HCl and NaOH have opposing [H +]/[OH ‐ ] concentrations.  The same result would apply when mixing equal volumes of a 6 pH acid and an 8 pH base, a 2 pH acid and a 12  pH base, and so on.       6  Chapter 3 - FUNDAMENTALS OF pH Definition of pH Just as the kilometer is a measure of distance and the hour is a measure of time, the pH unit measures the degree  of acidity or basicity of a solution.    To be more exact, pH is the measurement of the hydrogen ion concentration, [H+]. Every aqueous solution can be  measured to determine its pH value. This value ranges from 0 to 14 pH. Values below 7 pH exhibit acidic  properties. Values above 7 pH exhibit basic (also known as caustic or alkaline) properties. Since 7 pH is the center  of the measurement scale, it is neither acidic nor basic, it is called "neutral."    pH is defined as the negative logarithm of the hydrogen ion concentration. This definition of pH was introduced  in 1909 by the Danish biochemist, Soren Peter Lauritz Sorensen. It is expressed mathematically as:             where: [H+] is hydrogen ion concentration in mol/L  The pH value is an expression of the ratio of [H+] to [OH‐] (hydroxide ion concentration). Hence, if the [H+] is  greater than [OH‐], the solution is acidic. Conversely, if the [OH‐] is greater than the [H+], the solution is basic. At 7  pH, the ratio of [H+] to [OH‐] is equal and, therefore, the solution is neutral. As shown in the equation below, pH is  a logarithmic function. A change of one pH unit represents a 10‐fold change in concentration of hydrogen ion.  In a neutral solution, the [H+] = 1 x 10‐7 mol/L. This represents the pH of 7.                  Since the concentration of hydrogen ions and hydroxide ions is constant in a stable solution, either one can be  quantified if the value of the other is known. Therefore, when determining the pH of a solution, (even though the  hydrogen ion concentration is being measured), the hydroxide ion concentration can be calculated:          pH Values and Hydrogen/Hydroxide Concentrations   In Figure 3, the pH value corresponds to the number of decimal places under the column for "hydrogen ion  concentration." The pH of the solution equals the exponential form of the [H+], with the minus sign changed to a  plus. It is much easier to write or say "10 pH" than it is to communicate "a hydrogen‐ion concentration of  0.0000000001 mol/L."  7    How Is pH Measured? pH in an aqueous solution can be measured in a variety of ways. The most common way uses a pH‐sensitive glass  electrode, a reference electrode and a pH meter. Alternative methods for determining the pH of a solution are: • Indicators: Indicators are materials that are specifically designed to change color when exposed to different  pH values. The color of a wetted sample paper is matched to a color on a color chart to infer a pH value. pH  paper is available for narrow pH ranges (for example, 3.0 to 5.5 pH, 4.5 to 7.5 pH and 6.0 to 8.0 pH), and fairly  wide pH ranges of 1.0 to 11.0 pH.  NOTE:  pH  paper  is  typically  used  for  preliminary  and  small  volume  measuring.  It  cannot  be  used  for  continuous  monitoring  of  a  process.  Though  pH  paper  is  fairly  inexpensive,  it  can  be  attacked  by  process  solutions, which may interfere with the color change.  • Colorimeter: This device uses a vial filled with an appropriate volume of sample to which a reagent is added.  As the reagent is added, a color change takes place. The color of this solution is then compared to a color  wheel or spectral standard to interpolate the pH value.  The colorimeter can be used for grab sample measuring, but not for continuous on‐line measuring. Typically used  to determine the pH value of water in swimming pools, spas, cooling towers, and boilers, as well as lake and river  waters.  A pH meter is always recommended for precise and continuous measuring. Most laboratories use a pH meter  connected to a strip chart recorder or some other data acquisition device so that the reading can be recorded or  stored electronically over a user‐defined time range.    Activity versus Concentration Glass electrodes are sensitive to the hydrogen ion activity in a solution. Consequently, the concentration of  hydrogen ion is not the only factor influencing the pH of a solution. The concentration of other chemicals in the  solution, or the ionic strength of the solution, is also a major influence in the measurement of pH.    8  The term "ionic strength" describes the amount of ionic species in a solution, as well as the magnitude of charge  on those species. Examples of ion species compounds are sodium (Na+) sulfate (SO42‐), calcium (Ca2+) chloride (Cl‐ ), and potassium (K+) nitrate (NO3‐). Presence of these ions in solution tends to limit the mobility of the hydrogen  ion, thereby decreasing the activity of H+.    The concept of limited mobility of the hydrogen ion is analogous to a person entering a shopping mall. If the  shopping crowd is very small, the person is free to move about the mall in any direction. However, if the mall is  very crowded, the shopper has a difficult time moving from store to store, which severely limits their activity. It is  this same principle of a "crowded environment" that limits the activity of the hydrogen ion.  The following equation mathematically describes this effect on the activity of H+:      where:  f is the activity coefficient In solutions where the ionic strength is very low, the activity coefficient is 1.00, making the activity of hydrogen  ion equal to its concentration. As the ionic strength of a solution increases, the activity coefficient decreases.  This has the effect of lowering the activity of hydrogen ion, which is seen as an increase in pH. The following  example illustrates this point:    Example: The pH of a 0.00002 M solution of nitric acid can be calculated using this equation:            The value of [ f ] can be derived from various equations, or found in published tables:    CRC Handbook of Chemistry and Physics by Robert C. West, Ph.D., Ed., CRC Press, Inc., Boca Raton, FL    Lange’s Handbook of Chemistry by John A. Dean, Ph.D., Ed., McGraw‐Hill Book Company, NY, NY  The product of the activity coefficient and hydrogen ion concentration is equal to 0.00002. This means that the  ionic strength of the solution has no effect on the pH calculation.    If the ionic strength were 0.1, the new pH can be calculated using this equation:            If the ionic strength of the solution were 0.1, the activity coefficient, [ f ], would then be 0.75. The product of the  activity coefficient and hydrogen ion concentration is now less than 0.00002. This causes the pH calculation of  the nitric acid solution to increase by 0.12 pH unit. In this case, the ionic strength has a major influence on the  pH of the solution.  9  The Nernst Equation The general mathematical description of electrode behavior was described by the 19th century German chemist,  Hermann Walther Nernst (1864 – 1941). He introduced the Nernst equation in 1889, expressed as:     where:    E  =  total potential (in millivolts) between two electrodes   E0  =  standard potential of the ion      R  =  universal gas constant (in Joules/mol‐Kelvin)    T  =  absolute temperature (in Kelvin)    n  =  charge of the ion    F  =  Faraday constant (in Coulombs/mol)    ai  =  activity of the ion    The entire term "2.3RT/nF" is called the Nernst factor, or slope factor. This term provides the amount of change  in total potential for every ten‐fold change in ion concentration. For hydrogen ion activity, where n = 1, the  Nernst factor is 59.16 mV for every ten‐fold change in activity at 25°C. This means that for every pH unit change,  the total potential will change 59.16 mV.    The following general equation may be stated for any temperature (since pH is defined as the negative logarithm  of the hydrogen ion activity):      However, the Nernst factor will change when temperature changes; T is not constant. At 25⁰ C the slope of the  pH electrode is 59.16 mV/pH unit. At 0⁰ C the slope value is approximately 54 mV/pH, and at 100⁰ C the slope  value is approximately 74 mV/pH. The millivolt output of the glass pH electrode will change with temperature in  accordance with the Nernst equation. As the temperature increases, so does the millivolt output. Specifically, the  slope of the electrode is what changes.  The change in electrode output versus temperature is linear, which can be compensated in the pH meter. The  linear function for temperature vs. pH change can be expressed as:      If an uncompensated pH system were standardized in pH 7 buffer at 25⁰ C, and then a sample at 23⁰ C  measured 4.00 pH, the error would be 0.018 pH unit (.003 x 2⁰ C x 3 units). For a measurement of 4.00 pH at 75⁰  C (probably close to a typical worst case), an uncompensated pH system would read 4.45 pH.    The Standard Hydrogen Electrode The glass measuring electrode has its electrochemical roots planted in the earlier use of the standard hydrogen  electrode (SHE). The SHE is the universal reference for reporting relative half‐cell potentials. A type of gas  electrode, it was widely used in early studies as a reference electrode and as an indicator electrode for the  10  determination of pH values. The SHE could be used as either an anode or cathode depending upon the nature of  the half‐cell it is used with.  The SHE consists of a platinum electrode immersed in a solution with a hydrogen ion concentration of 1.00M.  The platinum electrode is made of a small square of platinum foil, which is platinized with a finely divided layer  of platinum (known as platinum black). Hydrogen gas, at a pressure of 1 atmosphere, is bubbled around the  platinum electrode. The platinum black serves as a large surface area for the reaction to take place, and the  stream of hydrogen keeps the solution saturated at the electrode site with respect to the gas.    It  is  interesting  to  note  that  even  though  the  SHE  is  the  universal  reference  standard,  it  exists  only  as  a  theoretical  electrode  that  scientists  use  as  the  definition  of  an  arbitrary  reference  electrode  with  a  half‐cell  potential of 0.00 volts. Because half‐cell potentials cannot be measured, this is the perfect electrode to allow  scientists  to  perform  theoretical  research  calculations.  The  reason  this  electrode  cannot  be  manufactured  because no solution can be prepared that yields a hydrogen ion activity of 1.00M.  11  Chapter 4 – THE pH SENSOR pH Electrodes A  pH  electrode  assembly,  or  sensor, consists of two primary parts: • Measuring electrode: The measuring electrode is sometimes called the glass electrode, and is also referred  to as a membrane or active electrode.  • Reference electrode: The reference electrode is also referred to as a standard electrode.    Figure 4 shows these two electrodes.    The pH measurement is comprised of two half‐cell, or electrode,  potentials. One half‐cell is the pH sensitive glass measuring electrode  and the other is the reference electrode. Just as the two half‐cell  potentials of a battery are required to complete a circuit so does a pH  sensor. The mathematical expression for this is:      where:  Em = the electrode potential of the measuring electrode    Er = the electrode potential of the reference  electrode    This type of measurement, in millivolts, is called a potentiometric   measurement. Since voltage‐measuring devices only determine differences in  potentials, there is no method for determining the potential of a  single electrode. A galvanic measurement circuit is formed by  connecting the measuring electrode (half‐cell potential) and the  reference electrode (half‐cell potential) to the signal input of the  measuring device. At the reference electrode, there is a  solid/solution interface, where a chemical reaction takes place. This  enables an electrical current to flow through the measuring device,  the pH meter, which allows the reading to be made (Figure 5).  Since the current that passes through the half‐cells and the solution  being measured is extremely small, the pH meter must have a high  internal impedance, so it does not "drag down" the millivolt  potential produced by the electrodes. This low current flow ensures  that the chemical characteristics of the solution being measured remains unaltered.    A galvanic potential is formed when charge exchanges occur at the phase boundaries of the glass measuring  electrode. In effect, the pH sensor assembly forms a galvanic cell using two metal conductors—lead wires of the  measuring and reference electrodes interconnected through their respective electrolyte solutions—and the  12  media. Since phase boundaries cannot be measured individually and  there are always more than two phase boundaries present, the pH  meter measures the overall potential. The overall potential is  comprised of the following elements:  • Metal lead‐out wire of the measuring electrode  • Electrolyte of the measuring electrode  • Diffusion potentials at solid/solution interfaces  • Electrolyte of the reference electrode  • Metal lead‐out wire of the reference electrode    The measuring and reference electrodes can be in one of two forms:  two physically separate electrodes, known as an electrode pair; or  the electrodes can be joined together in a single glass body  assembly known as a combination electrode (Figure 6).    The electrode pair and combination electrode styles originated many years ago and continue to be widely used  today. In the 1970s, a different style pH sensor was developed and patented specifically for continuous on‐line  measurement applications. This pH sensor uses a differential electrode technique (Figure 7) that employs two pH  glass measuring electrodes. One electrode is used as the active or measuring electrode, and the other is used as  part of a reference assembly. The reference assembly consists of a pH glass measuring electrode immersed in a 7  pH buffer solution which is mechanically isolated from the solution being measured by a double junction "salt  bridge" (see "The Reference Junction" subsection on page 14 for details). These two half‐cell potentials are then  referenced to a third ground electrode. The differential electrode technique is expressed as:      where:  Em = measuring electrode voltage  Er  = reference electrode assembly voltage   Eg = solution ground electrode voltage    After canceling the Eg term:        Other major components of the differential pH sensor include the  sensor cable, a temperature compensation device, electrolyte  solutions, and reference junctions. Based on field‐proven results, the differential electrode technique  has shown marked advantages over conventional electrode pairs  and combination electrodes. The double junction salt bridge (part  of the reference assembly) makes it extremely difficult for an  appreciable amount of the solution being measured to migrate into  the inner chamber. Since the inner chamber is filled with a buffer, a  100 to 1 dilution would only represent a change in measured pH of  0.05 pH units. Similar dilutions to the conventional reference  electrode used in electrode pairs or combination electrodes could  cause shifts of up to 2.0 pH units.  13  Another  advantage  of  the  differential  electrode  technique  is  the  third  "ground  electrode."  Since  ground  loop  currents  will  pass  through  only  this  electrode,  and  not  through  the  reference  electrode,  the  overall  pH  signal  output is unaffected by the ground loop potential.  The Measuring Electrode The galvanic voltage output produced by a measuring electrode will depend on the ionic activity of the species of  ions which the electrode was designed to measure. In the case of pH electrodes, it is the hydrogen ion activity. Based upon the Nernst equation, at 25° C, the output of a pH measuring electrode is equal to 59.16 mV per pH  unit. At 7.00 pH, which is the isopotential point for a perfect electrode, the output is 0 mV. As the solution pH  increases (less acidic), the mV potential becomes more negative. Conversely, as the solution pH decreases (more  acidic), the mV potential becomes more positive. The glass measuring electrode has been adopted as the measuring element for most pH sensors in use today. The  measurement is predicated on the principal that a hydrated gel layer forms between the outer surface of the  glass and the aqueous solution being measured (Figure 8).                                  The measuring electrode’s internal wire element has a potential, E3, with respect to the internal fill solution  (Figure 9). Another potential, E2, exists between the internal fill solution and the inside surface of the glass.    Depending on the pH of the solution being measured, hydrogen ions will migrate into or out of the gel layer. In an  alkaline solution, hydrogen ions migrate out of the gel layer and a negative charge is developed on the outer gel  layer. Because the electrode’s internal fill solution is at a constant pH value, the internal potential remains  constant. Therefore, the potential that is measured across the glass membrane is the result of the difference  between the inner and outer electrical charge.    Asymmetry Potential When a pH electrode is immersed in a solution with the same pH as its internal fill solution, there should not be a  measurable potential across the glass membrane. If such a potential exists, it is known as an asymmetry  potential (Figure 9). In practice, this potential is usually a few millivolts or less for a new, properly stored  electrode.   14  Factors that limit the ability of the ion exchange mechanism cause  asymmetry potential. This includes a dehydrated electrode, using the  electrode in a non‐aqueous solution, or plugging and/or coating of the  glass surface. Asymmetry potential is also referred to as the difference  in potentials between the measuring and reference electrodes when  immersed in a zero solution. This potential constantly changes  depending on the pH value of the solution, temperature, age of the  measuring electrode, and its type of glass formulation.    Calibration of the pH sensor, with the measuring instrument, will  compensate for any asymmetry potential(s). Proper cleaning of the  electrode and reference junction prior to calibration is essential. Long‐ term usage of the sensor between calibrations will show changing pH as  a function of the asymmetry potential(s) in combination with actual  changes of the pH value. Once the asymmetry potential has increased  beyond the instrument compensation capability through calibration,  the sensor must be renewed, if possible. If changing the internal fill  solution and/or junction does not reduce the asymmetry potential, the  sensor is beyond refurbishing and must be replaced.    Sodium Ion Error Although the pH glass measuring electrode responds very selectively to hydrogen ions, there is a small  interference caused by similar ions such as lithium, sodium, and potassium. The amount of this interference  decreases with increasing ion size. Since lithium ions are normally not in solutions, and potassium ions cause very  little interference, sodium ions present the most significant interference.   Sodium ion error, also referred to as alkaline error, is the result of alkali ions, particularly sodium ions,  penetrating the glass electrode silicon‐oxygen molecular structure and creating a potential difference between  the outer and inner surfaces of the electrode. Hydrogen ions are replaced with sodium ions, decreasing the  hydrogen ion activity, thereby artificially suppressing the true pH value. This is the reason pH is sometimes  referred to as a measure of the hydrogen ion activity and not hydrogen ion concentration.    Sodium ion interference occurs when the hydrogen ion concentration is very low and the sodium ion  concentration is very high. Temperature also directly affects this error. As the temperature of the process  increases, so does the sodium ion error.    Depending on the exact glass formulation, sodium ion interference may take effect at a higher or lower pH. There  is no glass formulation currently available that has zero sodium ion error. Since some error will always exist, it is  important that the error be consistent and repeatable. With many glass formulations, this is not possible since  the electrode becomes sensitized to the environment it was exposed to prior to experiencing high pH levels. For  example, the exact point at which the sodium ion error of an electrode occurs may be 11.50 pH, after immersion  in tap water, but 12.50 pH after immersion in an alkaline solution.    Controlled  molecular  etching  of  special  glass  formulations  can  keep  sodium  error  consistent  and  repeatable.  This  is  accomplished  by  stripping  away  one  molecular  layer  at  a  time.  This  special  characteristic  provides  a  consistent amount of lithium ions available for exchange with the hydrogen ions to produce a similar millivolt  potential for a similar condition.    15  Acid Error Acid error affects the low end of the pH measuring scale. As pH decreases and the acid error begins, water  activity is reduced due to higher concentrations of acid displacing water molecules. The thickness of the hydrated  gel layer becomes thinner due to acid stripping. This effect has a negative influence on the mV output, thereby  causing the measured pH value to remain higher than the theoretical pH value. The acid error changes very little  with temperature. Over time, an upward drift of pH in acidic solutions is indicative of acid error.    Since acid error is usually observed below 1.00 pH and most process applications are well above that, it is fairly  uncommon. Process monitoring equipment is usually set to respond to a setpoint at which appropriate action is  taken to return the pH to the setpoint value when the pH rises above or falls below it. For example, the  controller will add caustic to a solution that is below the setpoint. It does not matter that the true pH value is  0.91 pH and is not accurate, since the controller will be calling for caustic addition until the setpoint is reached.    Temperature Effects   Temperature affects the pH measurement in two ways. The first is a change in pH due to changes in dissociation  constants of the ions in the solution being measured. This implies that as solution’s temperature changes, the pH  value also changes. Presently available instrumentation cannot account for this change because the dissociation  constants vary from solution to solution.    The second reason temperature affects the pH measurement, is glass electrode resistance. Since the glass  measuring electrode is an ionic conductor, it stands to reason that the resistance of the glass will change as the  solution temperature changes. As temperature rises, resistance across the glass bulb decreases. This change in  resistance versus temperature is constant and can be calculated depending on the specific type of glass  formulation of the electrode.     In practical terms, electrode resistance drops ten‐fold for every 30° C rise in temperature. For example, an  electrode with a 100 megohm resistance at 25° C will decrease to 10 megohm at 55° C. Assuming that the  theoretical pH of the solution is constant, the changing electrode resistance will incorrectly affect the pH  reading, requiring the use of temperature compensation in the measurement circuit.     A typical glass electrode at 25° C can have a resistance of 100 megohms. Typical ranges can vary from 20 to 800  megohms. In contrast, the resistance range for reference electrodes varies from 100 to 5000 ohms.    The resistance value of the measuring electrode is based on  a number of influences: solution temperature, glass  formulation, glass thickness, the shape of the measuring  electrode tip (Figure 10), its surface measuring area, and the  physical state of the hydrated gel layer.    Although temperature will not affect the electrode’s  response speed, the thickness of the glass will. A thicker  glass will be more durable, but it also has more resistance, a  noisier signal and a longer response time. A thinner glass will  have less resistance and a quicker response time, but will be  more fragile.    16  With high electrode resistance comes the impact of capacitance  coupling and electrical noise (Figure 11). The lead wire exiting the  measuring electrode is prone to picking up spurious electrical  signals—an effect known as hand capacitance. Similar to  extending a radio antenna for pulling in distant stations, the  longer the electrode lead wire length, the more prone the  electrode is to picking up interference. In the case of hand  capacitance, merely waving a hand next to the cable will cause  the measuring instrument display to change erratically.    Shielding the electrode and using high quality cable in the  manufacturing process can minimize the effects of electrical  noise. Shielding is usually accomplished by incorporating a  metallic band within the glass measuring electrode body. This   band  extends  from  the  base  of  where  the  bulb  is coupled to  the electrode body, up the shaft, and back to where the cable  exits the body. When the electrode is immersed in the solution,  the band effectively shields the measuring element from stray  interference through the solution. Combination electrodes are inherently shielded by the reference fill solution  that surrounds the entire measuring element.    When taking measurements, it is best to immerse a majority of the metal band (the measuring electrode shaft)  into the solution. However, the area where the electrode shaft joins a cap or cable connection should not be  immersed. As for shielding the glass bulb, the solution being measured acts as the shield.    Most instruments presently manufactured compensate for electrode resistance changes resulting from solution  temperature fluctuations. This compensation is almost always accomplished automatically using a temperature‐ sensitive device as part of the measuring circuit. In other cases, it is accomplished manually with an adjustment  to the feedback circuit of the instrument’s electronics. On analog instruments, this is achieved by setting a dial  to the correct temperature setting. For microprocessor‐based instruments, a temperature value corresponding  to the actual temperature of the solution being measured is manually entered.    The Reference Electrode   The glass measuring electrode is significant, but only as part of the overall system used to measure pH. Its half‐ cell potential must be combined with the other half‐cell potential of the reference electrode to complete the  measurement circuit. Both of these elements, when submersed into a solution, generate a pH measurement  when connected to a measuring device. For the measuring electrode to provide an accurate measurement, the  reference electrode must have a constant and stable potential. Any deviation in its potential will cause the overall  potential to change, thereby causing the pH reading at the measuring device to also change.    The reference electrode consists of a silver wire coated with silver chloride that is immersed in an electrolyte  solution. This wire must be electrically connected with the solution being measured. This is accomplished through  a porous junction, commonly called a salt bridge, which physically isolates the electrolyte and wire from the  solution being measured. The electrolyte solution must have a high ionic strength to minimize resistance and not  affect the solution being measured, and must remain stable over large temperature swings.   Potassium   chloride    (KCl) solutions that are 3.0 molar, 3.5 molar, or saturated have been used successfully for many years. This  reference wire and electrolyte combination is the silver/silver chloride reference system. Other types of  reference systems are also manufactured for specific measurement needs. These alternate reference systems  consist of calomel, thalamid or mercury.  17  The Reference Junction     The reference junction is located at the measuring end of the reference electrode (or reference electrode  assembly for a differential pH sensor). The reference junction is also referred to as a "salt bridge," liquid junction,  or frit. Its purpose is to interface physically and electrically with the internal electrolyte and the solution being  measured. This reference junction completes the current path from the glass measuring electrode to the  reference electrode.    The reference junction must be chemically inert, so it does not to interfere with the ion exchange process but   allows small amounts of electrolyte to flow through it, while maintaining a consistent low resistance value. The  reference junction is constructed of porous materials such as wood, Teflon, Kynar, ceramic, or more exotic  materials such as asbestos or quartz fibers. There is also a ground glass sleeve junction that, just as the name  implies, uses two ground glass surfaces mated to one another. These two surfaces are fit tightly together, but  they allow electrolyte to permeate between them. The size of the junction material usually corresponds to the  size of the reference wire, and it is usually cylindrical.  Its diameter is typically  1/16 inch to 1/8 inch and its length  is between 1/8 inch and 1/2 inch. Annular junctions, which surround the reference electrode, are also used. They  are available in a wide range of materials and sizes.    For current to flow, the junction must be able to conduct electrons. This current flow is established by allowing  electrolyte solution (and to some detriment, the solution being measured) to penetrate the structure of the  junction. The junction is designed to enable small amounts of the reference electrode electrolyte to leach out, or  flow, through it into the solution being measured. As the electrolyte flows out through the junction, it prevents  the solution being measured from flowing back into the reference system and contaminating the KCl solution, or  attacking the silver and silver chloride (Ag/AgCl) wire.    However, there are times when the solution being measured does penetrate the junction and contaminate the  reference system. The solution being measured may be under sufficient pressure to force it back through the  junction into the reference system. With replaceable junctions, an incorrectly installed junction (missing O‐ring,  torn O‐ring, or cross threading) will allow the solution being measured to flow unimpeded past the junction.  Measuring a solution that tends to coat or plug the junction is also detrimental, since it stops the electrolyte  flow and increases the resistance of the junction.    Theoretically, the resistance of the junction and the chemical make‐up of the reference system are assumed to  be constant during calibration and measuring. However, due to junction contamination, junction plugging,  electrolyte dilution, and chemical attack of the silver/silver chloride wire, the resistance is always changing, as is   the chemical composition of the whole reference electrode. Consequently, this highlights the importance of  frequently calibrating to compensate for these factors.    Junction Potentials When comparing pH readings between two or more pH sensors, there are usually differences between the  readings. The comparisons may be between a process solution reading and a grab sample reading. In this case,  the differences are usually caused by the process reading being under pressure, while the laboratory reading is  not. Gases are normally entrained (dissolved) in the process solution, but dissipate from the grab sample and no  longer affect the pH reading. Additionally, the grab sample and process solution temperatures may be far enough  apart to affect the pH reading.     Using sensors of different types can also cause differences in readings.  Typically, combination electrodes or  differential pH sensors for process measurement and electrode pairs for laboratory measurement, or any  18  combination thereof can be the cause. The sensor for the process reading is usually a sealed device, whereas the  laboratory electrode pair is open to the atmosphere to enable refilling of electrolyte. Sealed references will  usually produce different zero points than those of laboratory refillable electrodes, resulting in different pH  readings.    Differences in readings may not necessarily be due to actual pH differences. Effects of junction potential,  electrical grounding, electrode membrane coating, and dissociation constants of the solution being measured can  greatly affect the millivolt output of the measuring and reference electrodes. Additionally, chemical compounds  or particulate matter in the solution can disable the reference electrode or junction.  The process can also  contaminate the silver/silver chloride reference system. Sulfur compounds, heavy metals, and strong oxidizing  and reducing agents can attack the silver/silver chloride reference system. The resultant reaction of this attack  will form silver complexes or precipitate elemental silver that can block the pores of the junction or produce  oxidation‐reduction potentials across the junction. In either case, this results in an offset reference potential. Gel-filled reference Electrode The gel‐filled reference electrode uses a gelling agent added to the reference electrolyte solution. The purpose is  to slow down the effects of contamination on the reference electrolyte by process migration through the porous  junction. It is also more difficult for this gelled electrolyte to be forced out of the reference system cavity while  the process undergoes periods of temperature and pressure cycling. The gel tends to last longer than  conventional electrolyte solution under normal circumstances. Higher temperatures and processes that attack or  strip the gelling agent when it is migrating through the porous junction will tend to break down the gelling agent. Buffers The term buffer is defined as "a substance that absorbs an impact, protects against a shock, resists change, or  maintains a relative acid/base concentration within a specific range."    With regard to pH, buffers are solutions in which the pH remains unchanged when small amounts of an acid or a  base are added. A buffer is a solution of a weak acid or base and one of its salts, that must resist a change in its  pH value. A buffered solution contains chemicals that will not readily allow a change in pH value when it is being  neutralized or changed by dilution.    An example of a poor buffer solution is trisodium phosphate. A 0.06% solution at room temperature has a pH of  12.0. A solution with half as much concentration (0.03%) has a pH of about 11.7, while a solution with twice as  much concentration (1.2%) has a pH of 12.5. It is evident that an increase or decrease in solution concentration  will also increase or decrease the pH.    An example of a good buffer is a solution of sodium sesquicarbonate. The pH value for this solution, in the same  concentrations and with the same temperature as the previous example, remains consistent. This buffer  indicates that it is a more stable solution.  Calibration pH electrodes must be calibrated from time to time to maintain measurement accuracy. It is a fact that the  performance of a pH sensor degrades over time. The time period and related loss of sensor performance varies  considerably with each application and its unique conditions.    19  Specifically, calibration is performed to compensate for changes in potential within the measuring and reference  electrodes, as well as any change of potential between them. The electrodes are usually matched at the factory  so that the measuring electrode and reference electrode, when put in a zero solution, (7.0 pH buffer) provide a  zero mV output.    Differences or changes in potential can be caused by one or more of the following factors:    • Contamination of the reference electrolyte solution. The solution being measured is allowed to migrate  through the junction, entering the electrolyte cavity, thereby changing its composition and altering its  electrochemical reaction    • Electrolyte evaporation or depletion. As the solution being measured enters the reference electrolyte  cavity, the electrolyte is displaced. Under the right combination of temperature and pressure conditions,  the electrolyte, process solution, or a mixture of both can be drawn out of the cavity. As the level changes,  so does the potential output, until finally, the level drops below the silver/silver chloride wire to cause an  open measuring circuit.    • Chemical attack of the silver/silver chloride wire. As the solution being measured enters the reference  electrolyte cavity, certain changes may occur that poison the silver/silver chloride wire. This physical and  chemical attack on the wire changes its properties so that the whole reference system no longer produces  the same output as it did prior to the attack. In some cases, this change in output will remain constant,  while in others it may be constantly degenerating over time.    • Junction potential. When exposed to the solution being measured, the reference junction may become  contaminated or plugged. This changes the resistance of the reference electrode, thereby changing its  output.    • Aging of the measuring electrode. As the glass measuring electrode is exposed to the solution being  measured, the electrode is continuously attacked. The gel layer that is formed at the electrode tip undergoes  continuous change, which ultimately alters its output. This "aging" is continuous. The glass measuring  electrode never becomes stable. As aging progresses, the gel layer becomes thicker, thereby affecting its  output.    Taking into account any one of these factors, let alone the possibility of two or more of these conditions  occurring, it is easy to see why a pH sensor must be recalibrated in a pH buffer solution to maintain measurement  accuracy. Calibration also involves checking the slope of the measuring electrode. Slope defines the ability of the measuring  electrode to change its output by 59.16 mV per pH unit at 25° C. For previously stated reasons, virtually all pH  instruments use a slope adjustment to compensate for the measuring electrode’s inability to accurately produce  its output signal. The slope adjustment is made using a buffer that has a value at least 3 pH units from the zero  buffer (7.0 pH).    Buffers are used when calibrating pH sensors to compensate for any inherent offsets of the sensor. Typically, 4  pH and 7 pH buffers are used to perform the calibration. For best calibration accuracy, use buffers with values  that are close to the normal pH value of the solution being measured. For example, if the solution being  measured is normally 2 pH, it is best to use 4 pH and 7 pH buffers rather than 7 pH and 10 pH buffers. Ideally, one  of the buffer values for this example should be 2 pH.    20  Adhere to the standard practice of using calibration buffers that are at least three  pH units apart. This difference   is intended to provide the electrode with sufficient change, enabling its slope to be plotted over a wide area.  When adjusting the zero and slope, carefully follow the equipment manufacturer’s instructions to avoid  inaccurate results. When calibrating an instrument without temperature compensation capability, the  temperatures of the buffers need to be known.     Figure 12 indicates how the value of a pH buffer is affected by  changes in temperature.    For instruments with automatic temperature compensation, the  proper calibration technique is to allow the sensor and buffer  temperatures to equalize, and to rinse the sensor in distilled  water between buffers. This will remove residual contaminants  from the sensor, prevent buffer contamination carryover, and  prolong the buffer solution. Never reuse buffers by pouring the  used buffer portion back into the bottles. Always discard buffers  immediately after use.    The following graphs show the temperature versus pH curve in a more representative form:          21          NOTE: pH 10 buffer is also readily available, but it is not as stable as pH 7 and pH 4 buffers, particularly at  extreme temperatures. When pH 10 buffer is exposed to air, it absorbs carbon dioxide. As this occurs, the buffer  becomes more acidic and is no longer dependable as a calibration reference solution. Even when stored in  sealed plastic bottles, the carbon dioxide molecules permeate the plastic bottle and over time, this will cause  the same reaction.      22  Chapter 5 – CARE OF pH ELECTRODES Dehydration Left out of solution, the pH glass membrane will become dehydrated. After this happens, the pH sensor will have  slower response and a higher than normal impedance when it is put back into operation. Repeated dehydration   and re‐use will dramatically reduce the normal service life of the pH sensor. Prolonged dehydration will cause the  glass membrane to fail completely.  If the reference electrode becomes dehydrated, it will no longer operate properly. The electrolyte will leach out  of the electrode cavity, through the junction(s), forming salt crystals on the junction surface. Over time, leaching  will weaken the electrolyte potential, and may also cause a phenomenon known as a bridging effect. Both of  these conditions will increase the output impedance, making the reference electrode output unstable. With  continued dehydration, the impedance will rise to a level that becomes unusable to the pH meter.  Factors Detrimental to Electrode Life A pH electrode operates similarly to a hydrogen electrode within the range of 0.00 ‐ 12.00 pH (where the alkali  error affects the reading). This is known as a sodium ion error (discussed earlier in Chapter 4 – The pH Sensor).  Within this range, the output slope of the electrode corresponds to the theoretical 59.16 mV as defined by the  Nernst equation.    As with all glass, pH glass is susceptible to chemical attack. Temperature changes can alter the rate of this attack.  For every 30° C rise in temperature, the rate of attack increases ten‐fold. Accordingly, process solutions with  elevated temperatures shorten electrode life. Strong acids and strong alkaline solutions, to a  greater extent,  attack the glass membrane. Even neutral solutions that contain high concentrations of alkali ions, sodium ions in  particular, attack the glass. Using a pH sensor with a glass formulation that is inappropriate for the application  may render the sensor inoperable after only a short time without any visible glass defects.    Hydrofluoric acid (HF) will readily poison the glass membrane when the pH is below 6.00. The greater the fluoride  ion concentration, the faster the electrode will fail. The fluoride strips away the gel layer of the glass membrane  rendering it inoperative.    A special electrode manufactured from antimony is available for measuring pH in solutions containing HF. The  antimony electrode exhibits similar properties to glass electrodes within certain limits. One drawback is that the  repeatability and speed of response for an antimony electrode is inferior to that of a glass electrode. Also,  antimony electrodes are only linear between 3.0 and 8.0 pH, and should only be specified when the presence of  hydrofluoric acid dictates their use.    Transportation Freezing, extreme heat, vibration, and mechanical shock must be avoided when transporting electrodes, whether  within a facility, or from one facility to another. Always try to reuse the original box and packing materials to  transport electrodes.    When shipping the electrode using motor freight, select a carrier that will guarantee that the package will not be  exposed to extreme temperatures. Usually sending the electrode by an overnight delivery service ensures that  the package will not be exposed to the elements long enough to damage the electrode.  23  Storage pH sensors (electrode pairs, combination electrodes, and differential styles) should be stored in ambient  conditions between 10 and 30⁰ C. Protective caps and solution storage caps should be kept intact and installed on  the end of the sensor as provided by the manufacturer.    The best solution for storage purposes is a 3 to 3.5 M KCl solution. This solution provides a neutral to slightly   acidic environment for the glass electrode, and will not impose a memory on the glass, much as Ni‐Cad batteries  can have memories imposed upon them when they are not fully discharged prior to recharging. Should KCl  solution be unavailable, appropriate substitutes in order of preference are:    1. pH 4 buffer   2. Distilled water  3. Tap water    Under these conditions, the glass measuring and reference electrodes can be stored for three to five years.    NOTE: Periodically check to verify that the storage solution has not evaporated.      24  Chapter 6 – COMMON APPLICATIONS Cooling Tower Control The purpose of a cooling tower is to cool down industrial processes and provide cooled water for HVAC control.  The cooled process water is directed through the process, where it absorbs heat, and is then sent back to the  tower to be re‐cooled. Large amounts of water are required as evaporation occurs during the cooling cycle. Since  water is an expensive commodity, primary concerns are efficiencies in cooling tower design and chemical  treatment of the water being used.    Hot water entering the cooling tower is sprayed through cooler air to speed evaporation. For efficient cooling, it  is very important to maintain a large surface area for exposing the water to the air for an extended period of  time. These heat transfer surfaces must be protected from corrosion and scale. After the water is cooled, it is  collected in a sump and pumped back through the process piping.    Proper control of a number of parameters will keep the tower operating efficiently and prevent damage to vital  parts from scale, corrosion, and biological growth. As air flows through the tower, airborne contaminants are  picked up by the water/air interface. These contaminants would be carried through the system if they were not  controlled prior to distribution.    pH is one of many parameters that controls the water chemistry of the tower. Maintaining a pH level between  6.0 and 7.0 is fairly common. Depending on the pH of the water supply to the tower—lake, stream, or municipal  water supply—either an acid or caustic control scheme is used to maintain the pH in this range.  Food Processing Canning plants use caustic soda (NaOH) in their fruit and vegetable peeling operations. Prior to disposal of the  wastes, the pH must be neutralized. Usually this is accomplished using CO2 (carbon dioxide) as a reagent. The CO2  is injected into stainless steel pipelines to lower the caustic waste, typically from a pH of 12.7 to 9.5. The amount  of CO2 reagent added to the peeling wastes is determined by the resultant pH value of the waste stream at the  plant discharge point.    The neutralized waste material is recycled and used as both wet and dry animal feed for surrounding farms.    Coal Industry Water run‐off from coal mining operations is acidic and must be  collected and  treated prior to  contact with   surrounding water sheds. Anhydrous ammonia is typically used as a reagent to increase the pH. Values of coal  run‐off water have been known to be as low as 2.05 pH.     Under atmospheric conditions, anhydrous ammonia is a vapor. Stored in a tank under pressure, it is a liquid.  Specially engineered systems can feed either liquid or vaporous anhydrous ammonia. Depending on the system  design criteria, each form has its particular advantages and disadvantages. When feeding anhydrous ammonia  in vapor form, an electrically heated vaporizer must be used. Ammonia is easier to measure and control if  applied as a vapor.    25  Run‐off water is collected or pooled. The pH of this effluent is measured and then compared to the control  setpoint. If the pH is below the acceptable level, the pH controller controls a valve to allow the reagent to be  added to the effluent and mixed until the optimum pH value is reached. After the collected water is  appropriately treated, it is discharged.    Plating Waste Treatment The plating industry produces wastes of cyanides, chromates, acids, and alkali cleaners. Each must be treated  prior to discharge. Typically, pH and Oxidation Reduction Potential (ORP) measurement and control are used in  combination to neutralize and detoxify these solutions. The majority of applications involve batch treatment in  holding tanks with good mixing.    Treating toxic solutions before they become part of the effluent flow is preferable to final stage treatment.  Cyanide and chromate wastes are just two examples that use this scheme. In this scenario, a tank with the  treatment chemicals is downstream from the plating tank. Rather than being rinsed from the parts and treated  later, carry over chemicals are treated within this tank.    Chlorine or sodium hypochlorite is used to oxidize cyanide to a less toxic cyanate. Proper care must be taken to  maintain an alkaline pH so that the pH of the solution does not fall, producing deadly cyanide gas. Caustic is  added to the process through a solenoid valve controlled by the pH measurement system. The pH during this  treatment phase must be closely maintained between 9.98 and 10.02 pH. The reaction is furthered when the ORP  value, monitored by an ORP measurement system, reaches a pre‐determined point for a specified time.    The final step involves either pumping the solution to another tank to extract the heavy metals, or adding acid to  the solution to get a final 7.5 to 8.0 pH level. This process starts the oxidation of the cyanate. After proper pH  adjustment, the cyanate is oxidized to carbon dioxide, nitrogen, and water by adding chlorine or hypochlorite.  Again, proper monitoring of ORP during this step ensures a completed reaction.    Ultrapure Water The measurement of pH in ultrapure water requires some special considerations. Because of its unique  properties, ultrapure water has been called the world’s greatest solvent. Its purity and solvent properties prevent  corrosion of wetted parts in boilers in power generating plants. It is used as a cleaner to wash contaminants from  electronic microchips in the electronics industry, and it provides the base for injectable medicines in the  pharmaceutical industry.    Chemically pure water consists of two atoms of hydrogen and one atom of oxygen. It is odorless and colorless.  Italian scientist Stanislao Cannizzarro defined the chemical formula of the water molecule, H20, in 1860.  Absolutely pure water does not last long because it dissolves nearly everything it comes into contact with,  thereby retaining some of the properties of the contacted material. It readily absorbs carbon dioxide when  exposed to the atmosphere. It is very difficult to make pure water even through repeated distillations.    When heated, water boils producing steam (water vapor) as a byproduct. Considered pure, steam quickly absorbs  atmospheric elements such as carbon dioxide that quickly change the water to something other than ultrapure  water. Because ultrapure water has no buffer capacity, the slightest contamination will change its pH value.    Pure water is a very good electrical insulator. This makes it especially difficult to conduct electrons between the  measuring and reference electrodes of a pH sensor. Within the pH electrode measuring circuit, this causes a  high resistance that is prone to electrical noise, slow electrode response, hand capacitance effects, and static  26  buildup. This static buildup is even more prevalent when pure water flows through plastic piping. A common  phenomenon stemming from this buildup of static charge is referred to as a streaming current potential. This  potential will affect the pH reading by pro‐ viding false electrical potentials to the measuring and reference  electrodes. These potentials can cause a variable or constant voltage offset to be applied to the electrodes,  making the pH measurement meaningless. After repeated  exposure to ultrapure water, the reference junction  will exhibit a high resistance due to its insulating properties, causing drift and unstable calibration.    Electrodes for measuring ultrapure water are specialized and are usually mated with a metal, grounded flow  assembly. The metal flow assembly helps shield the reading from electrical noise and hand capacitance. A slow  process flow rate through the assembly also provides ample residence time for the measurement to occur and,  to some degree, eliminates streaming current potentials.    The electrode  assemblies  have gel‐filled  references or flowing junctions so a fast and efficient potential is  established between the junction and the solution being measured. The electrolyte passing through the junction  provides this potential with a minimal chance of junction plugging or fouling. The only drawback of this design  may be the contamination of the process solution with the electrolyte. Putting the entire measuring loop into a  bypass assembly and then directing the sample discharge to drain will keep a constant pressure on the measuring  electrode. This will also negate the problem of sending a process sample, contaminated with KCl electrolyte, back  into the process.      27  BIBLIOGRAPHY   Arthur, Robert M. 1982. Application of On‐line Analytical Instrumentation to Process Control. Arthur  Technology. Bates, Roger G. 1973. Determination of pH, Theory and Practice.  Beckman 1983. The Beckman Handbook of Applied Electrochemistry.  Beckman Instruments, Inc.   Chemtrix. pH in Plain Language. Chemtrix, Inc.  Gray, David M. 1994. On‐line High Purity pH Measurement. Leeds and Northrup.  Ingold, 1989. Practice and Theory of pH Measurement. Ingold Messtechnik AG.  McMillan, Gregory K. 1994. pH Measurement and Control; Second Edition. Instrument Society of America.  Merriman, Dale C. 1993. Reference Junction Potential Differences Between Laboratory and Process  Instrumentation. Instrument Society of America.  Moore, Ralph L. 1978. Neutralization of Waste Water by pH Control. Instrument Society of America.   Schott Gerate. Electrodes and Accessories for Technical Service. Schott Corporation.      28  GLOSSARY Activity: A thermodynamic term for the apparent or active concentration of a free ion in solution. It is related to  con‐ centration by the activity coefficient.  Asymmetry Potential: The potential developed across the glass membrane of the measuring electrode with  identical solutions on both sides. Also a term used when comparing measuring electrode potential in pH 7 buffer.  ATC:  Automatic temperature compensation.  BNC: A quick disconnect electrical connector used to inter‐ connect and/or terminate coaxial cables.  Buffer Capacity: A measure of the ability of the solution to resist pH change when a strong acid or base is added.  Buffer: Any substance or combination of substances which, when dissolved in water, produces a solution that  resists a change in its hydrogen ion concentration when acid or alkali is added.  Calibration: The process of adjusting an instrument or compiling a deviation chart so that its reading can be  correlated to the actual value being measured.  Conductance: The measure of the ability of a solution to carry an electrical current.  Dissociation Constant: A value which quantitatively expresses the extent to which substances dissociate in  solution. The smaller the value of K is, the less dissociation of the species in solution. This value varies with  temperature, ionic strength, and the nature of the solvent.  Drift: A change of a reading or a setpoint value over a period of time due to several factors including change in  ambient temperature, time, and the line voltage.  Electrode Potential: The difference in potential established between an electrode and a solution in which it is  immersed.  Electrolyte:  Any substance in a solution that will conduct an electric current. Acids, bases, and salts are common  electrolytes.  Filling Solution:  A solution of defined composition, also called an electrolyte, that provides a chemical reaction  as well as an electrical potential between an internal element and the process being measured. An example is  the solution sealed inside a pH glass bulb, typically KCl. A filling solution normally is a buffered chloride, which  provides a stable potential and a specific zero potential point. In a reference electrode, the electrolyte, also  called the reference filling solution, surrounds the silver/silver chloride wire and periodically requires  replenishing.  Hydrogen Ion Activity:  Activity of the hydrogen ion in solution. It is related to hydrogen ion concentration (CH+)  by the activity coefficient for hydrogen (ƒH+).  Impedance:  The total opposition (resistive plus reactive) to electrical flow.  Input Resistance (Impedance): The input resistance of a pH meter is the resistance between the measuring  electrode terminal and the reference electrode terminal. A voltage division between the total electrode  resistance and the input resistance always affects the potential of a pH measuring electrode circuit.  Internal Reference Electrode Wire: The silver/silver chloride wire used in a reference electrode.    Isopotential Point: A potential which is not affected by temperature changes. It is the pH value at which dE/dt  for a given measuring/reference electrode system is zero. Theoretically, for a glass measuring electrode and SHE  reference electrode, this potential exists when immersed in pH 7 buffer.  29  Logarithmic Scale: A method of displaying data (in powers of ten) to yield maximum range while keeping  resolution at the low end of the scale.  Membrane: The pH sensitive glass bulb is the membrane across which a potential difference is developed from  an ion‐ exchange reaction. The membrane separates the electrolyte in the measuring electrode from the  solution being measured.  Millivolt (mV): A unit of electromotive force. It is the difference in  potential  required  to  make  a  current  of 1  millampere flow through a resistance of 1 ohm. One millivolt equals one thousandth of a volt.  Molality: A measure of concentration expressed in moles per kilogram of solvent.  Molarity: A measure of concentration expressed in moles per liter of solution.  Nernst Equation: A mathematical description of electrode behavior in which: E is the total potential, in millivolts,  developed between the measuring and reference electrodes; Ex varies with the choice of electrodes,  temperature, and pressure; 2.3RT/nF is the Nernst factor (R and F are constants, n is the charge on the ion  including sign, and T is the temperature in degrees Kelvin); and ai is the activity of the ion to which the electrode  is responding.  Nernst Factor (Slope): The term 2.3RT/nF in the Nernst equation, when T = 25° C, is equal to 59.16 mV when n =  1, and 29.58 mV when n = 2. The term n is the sign of the charge on the ion. The Nernst factor varies with  temperature.  pH Junctions: The junction of a reference electrode or combination electrode, typically a liquid junction, is a  permeable membrane through which the electrolyte migrates.  Salt Bridge: The salt bridge of a reference electrode is that part of the electrode which is in contact with the  electrolyte, establishing the electrolytic connection between the reference  system  and  the   solution   being    measured. An "auxiliary" salt bridge is a glass tube open at one end for electrolyte filling, and connected to the  reference electrode cavity at the other end. This type of salt bridge, used for special applications, increases the  electrolyte capacity.  Span:  The difference between the upper and lower limits of a range expressed in the same units as the range.  Stability: The quality of an instrument or sensor to maintain a consistent output when a constant input is  applied.  Standardization: A process, also known as calibration, of equalizing the measuring electrode and reference  electrode millivolt output potentials in one standardizing solution (buffer) so that potentials developed in  unknown solutions can be interpreted as pH values.  Thermistor: A temperature‐sensing element composed of sintered semiconductor material which exhibits a  large change in resistance proportional to a small change in temperature. Thermistors usually have negative  temperature coefficients.                LITG004  © Hach Company, 2010.  All rights reserved.  In the interest of improving and updating its equipment, Hach Company reserves the right to alter specifications to equipment at any time.