Teledyne Analytcial Instruments OPERATING INSTRUCTIONS FOR MODEL 3000TA-EU Trace Oxygen Analyzer
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
OPERATING INSTRUCTIONS FOR
MODEL 3000TA-EU
Trace Oxygen Analyzer
P/N M66316
4/15/16
DANGER
Toxic gases and or flammable liquids may be present in this monitoring
system.
Personal protective equipment may be required when servicing this
instrument.
Hazardous voltages exist on certain components internally which may persist
for a time even after the power is turned off and disconnected.
Only authorized personnel should conduct maintenance and/or servicing.
Before conducting any maintenance or servicing, consult with authorized
supervisor/manager.
Teledyne Analytical Instruments
3000TA- EU
Copyright © 2016 Teledyne Analytical Instruments
All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed,
stored in a retrieval system, or translated into any other language or computer language in
whole or in part, in any form or by any means, whether it be electronic, mechanical,
magnetic, optical, manual, or otherwise, without the prior written consent of Teledyne
Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91748.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us free from
defects of material and of construction, and that our liability shall be limited to replacing or
repairing at our factory (without charge, except for transportation), or at customer plant at our
option, any material or construction in which defects become apparent within one year from the
date of shipment, except in cases where quotations or acknowledgements provide for a shorter
period. Components manufactured by others bear the warranty of their manufacturer. This
warranty does not cover defects caused by wear, accident, misuse, neglect or repairs other than
those performed by Teledyne or an authorized service center. We assume no liability for direct
or indirect damages of any kind and the purchaser by the acceptance of the equipment will
assume all liability for any damage which may result from its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our apparatus,
and to make any alterations in the dimensions, shape or weight of any parts, in so far as
such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its user, and serves as a tool by which
valuable data can be gathered. The information provided by the instrument may assist the
user in eliminating potential hazards caused by his process; however, it is essential that all
personnel involved in the use of the instrument or its interface, with the process being
measured, be properly trained in the process itself, as well as all instrumentation related to
it.
The safety of personnel is ultimately the responsibility of those who control process
conditions. While this instrument may be able to provide early warning of imminent
danger, it has no control over process conditions, and it can be misused. In particular, any
alarm or control systems installed must be tested and understood, both as to how they
operate and as to how they can be defeated. Any safeguards required such as locks, labels,
or redundancy, must be provided by the user or specifically requested of Teledyne at the
time the order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions. The purchaser
is responsible for the training of personnel, for providing hazard warning methods and
instrumentation per the appropriate standards, and for ensuring that hazard warning devices
and instrumentation are maintained and operated properly.
Teledyne Analytical Instruments, the manufacturer of this instrument, cannot accept
responsibility for conditions beyond its knowledge and control. No statement expressed or
implied by this document or any information disseminated by the manufacturer or its
agents, is to be construed as a warranty of adequate safety control under the user’s process
conditions.
Teledyne Analytical Instruments
ii
Trace Oxygen Analyzer
Specific Model Information
The instrument for which this manual was supplied may
incorporate one or more options not supplied in the standard instrument.
Commonly available options are listed below, with check boxes. Any
that are incorporated in the instrument for which this manual is supplied
are indicated by a check mark in the box.
Instrument Serial Number: _______________________
Options Included in the Instrument with the Above Serial Number:
3000TA-EU-C: In addition to all standard features, this model
also has separate ports for zero and span gases,
and built-in control valves. The internal valves
are entirely under the control of the 3000TAEU electronics to automatically switch between
gases in synchronization with the analyzer’s
operations
19" Rack Mnt: The 19" Relay Rack Mount units are available
with either one or two 3000 series analyzers
installed in a standard 19" panel and ready to
mount in a standard instrument rack.
Sensor Options Available for the Instrument with the Above Serial
Number:
Insta-Trace B2C
A2C
B2C
L2C
Insta-Trace A2C
Teledyne Analytical Instruments
iii
3000TA- EU
Important Notice
Model 3000TA-EU complies with all of the requirements of the
Commonwealth of Europe (CE) for Radio Frequency Interference,
Electromagnetic Interference (RFI/EMI), and Low Voltage Directive
(LVD).
The following International Symbols are used throughout the
Instruction Manual. These symbols are visual indicators of important
and immediate warnings and when you must exercise CAUTION while
operating the instrument. See also the Safety Information on the next
page.
STAND-BY: Instrument is on Stand-by, but circuit is
active
GROUND: Protective Earth
CAUTION: The operator needs to refer to the manual
for further information. Failure to do so may
compromise the safe operation of the equipment.
CAUTION: Risk of Electrical Shock
Teledyne Analytical Instruments
iv
Trace Oxygen Analyzer
Safety Messages
Your safety and the safety of others is very important. We have
provided many important safety messages in this manual. Please read
these messages carefully.
A safety message alerts you to potential hazards that could hurt you
or others. Each safety message is associated with a safety alert symbol.
These symbols are found in the manual and inside the instrument. The
definition of these symbols is described below:
No
Symbol
GENERAL WARNING/CAUTION: Refer to the
instructions for details on the specific danger. These cautions
warn of specific procedures which if not followed could
cause bodily Injury and/or damage the instrument.
CAUTION: HOT SURFACE W ARNING: This warning is
specific to heated components within the instrument. Failure
to heed the warning could result in serious burns to skin and
underlying tissue.
WARNING: ELECTRICAL SHOCK HAZARD: Dangerous
voltages appear within this instrument. This warning is
specific to an electrical hazard existing at or nearby the
component or procedure under discussion. Failure to heed
this warning could result in injury and/or death from
electrocution.
Technician Symbol: All operations marked with this
symbol are to be performed by qualified maintenance
personnel only.
NOTE: Additional information and comments regarding a
specific component or procedure are highlighted in the form
of a note.
Teledyne Analytical Instruments
v
3000TA- EU
CAUTION:
THE ANALYZER SHOULD ONLY BE USED FOR THE
PURPOSE AND IN THE MANNER DESCRIBED IN
THIS MANUAL.
IF YOU USE THE ANALYZER IN A MANNER OTHER
THAN THAT FOR WHICH IT WAS INTENDED,
UNPREDICTABLE BEHAVIOR COULD RESULT
POSSIBLY ACCOMPANIED WITH HAZARDOUS
CONSEQUENCES.
This manual provides information designed to guide you through
the installation, calibration and operation of your new analyzer. Please
read this manual and keep it available.
Occasionally, some instruments are customized for a particular
application or features and/or options added per customer requests.
Please check the front of this manual for any additional information in
the form of an Addendum which discusses specific information,
procedures, cautions and warnings that may be peculiar to your
instrument.
Manuals do get lost. Additional manuals can be obtained from
Teledyne at the address given in the Appendix. Some of our manuals are
available in electronic form via the internet. Please visit our website at:
www.teledyne-ai.com.
Teledyne Analytical Instruments
vi
Trace Oxygen Analyzer
This is a general purpose instrument designed for use in a nonhazardous
area. It is the customer's responsibility to ensure safety especially when
combustible gases are being analyzed since the potential of gas leaks
always exist.
The customer should ensure that the principles of operation of this
equipment are well understood by the user. Misuse of this product in
any manner, tampering with its components, or unauthorized
substitution of any component may adversely affect the safety of this
instrument.
Since the use of this instrument is beyond the control of Teledyne, no
responsibility by Teledyne, its affiliates, and agents for damage or injury
from misuse or neglect of this equipment is implied or assumed.
Teledyne Analytical Instruments
vii
3000TA- EU
Table of Contents
Safety Messages ........................................................................... v
Introduction ................................................................................... 1
1.1 Overview
1
1.2 Typical Applications
1
1.3 Main Features of the Analyzer
1
1.4 Model Designations
2
1.5 Front Panel (Operator Interface)
3
1.6 Recognizing Difference Between LCD & VFD
4
1.7 Rear Panel (Equipment Interface)
5
Operational Theory ....................................................................... 7
2.1 Introduction
7
2.2 Micro-Fuel Cell Sensor
7
2.2.1 Principles of Operation
7
2.2.2 Anatomy of a Micro-Fuel Cell
8
2.2.3 Electrochemical Reactions
9
2.2.4 The Effect of Pressure
10
2.2.5 Calibration Characteristics
10
2.3 Sample System
11
2.4 Electronics and Signal Processing
13
Installation ................................................................................... 17
3.1 Unpacking the Analyzer
17
3.2 Mounting the Analyzer
17
3.3 Rear Panel Connections
19
3.3.1 Gas Connections
19
3.3.2 Electrical Connections
21
3.3.2.1 Primary Input Power
21
3.3.2.2 50-Pin Equipment Interface Connector
22
3.3.2.3
RS-232 Port
Teledyne Analytical Instruments
viii
27
Trace Oxygen Analyzer
3.4 Installing the Micro-Fuel Cell
3.5 Testing the System
29
29
Operation ..................................................................................... 31
4.1 Introduction
31
4.2 Using the Data Entry and Function Buttons
31
4.3 The System Function
33
4.3.1 Tracking Oxygen Readings During Calibration and
Alarm Delay
34
4.3.2 Setting up an Auto-Cal
36
4.3.3 Password Protection
37
4.3.3.1 Entering the Password
37
4.3.3.2 Installing or Changing the Password
38
4.3.4 Logout
40
4.3.5 System Self-Diagnostic Test
41
4.3.6 Version Screen
42
4.3.7 Showing Negative Oxygen Readings
42
4.4 The Zero and Span Functions
43
4.4.1 Zero Cal
44
4.4.1.1 Auto Mode Zeroing
44
4.4.1.2 Manual Mode Zeroing
45
4.4.1.3 Cell Failure
45
4.4.2 Span Cal
46
4.4.2.1 Auto Mode Spanning
46
4.4.2.2 Manual Mode Spanning
47
4.4.3 Span Failure
49
4.5 The Alarms Function
49
4.6 The Range Function
51
4.6.1 Setting the Analog Output Ranges
52
4.6.2 Fixed Range Analysis
53
4.7 The Analyze Function
54
4.8 Signal Output
54
4.9 Maintenance Schedule
55
4.10 Sensor Detection
57
4.11 Valve Box Functions
59
Teledyne Analytical Instruments
ix
3000TA- EU
4.11.1 Serial#
4.11.2 Stream
4.11.3 SerMode
59
60
61
Maintenance ................................................................................ 64
5.1 Routine Maintenance
64
5.2 Cell Replacement
64
5.2.1 Storing and Handling Replacement Cells
64
5.2.2 When to Replace a Cell
65
5.2.3 Removing the Micro-Fuel Cell
65
5.2.4 Installing a New Micro-Fuel Cell
67
5.2.4.1 Standard Trace Oxygen Sensor Cell Installation
67
5.2.4.2 Insta-Trace Cell Installation
68
5.3 Fuse Replacement
69
5.4 System Self Diagnostic Test
70
5.5 Major Internal Components
71
5.6 Cleaning
72
5.7 Troubleshooting
72
Appendix ...................................................................................... 74
A-1 Model 3000TA-EU Specifications
74
A-2 Recommended 2-Year Spare Parts List
75
A-3 Drawing List
77
A-4 19-inch Relay Rack Panel Mount
77
A.5 Application notes
78
A-5 Material Safety Data Sheet
82
Teledyne Analytical Instruments
x
Trace Oxygen Analyzer
List of Figures
Figure 1-1: Model 3000TA-EU Front Panel ..................................... 3
Figure 1-2: Model 3000 TA Rear Panel ........................................... 5
Figure 2-1: Micro-Fuel Cell .............................................................. 8
Figure 2-2. Cross Section of a Micro-Fuel Cell (not to scale) .......... 8
Figure 2-3. Characteristic Input/Output Curve for a Micro-Fuel
Cell ............................................................................. 11
Figure 2-4: Piping Layout and Flow Diagram for Standard Model . 12
Figure 2-5: Flow Diagram .............................................................. 13
Figure 2-6: 3000TA-EU Internal Electronic Component Location .. 14
Figure 2-7: Block Diagram of the Model 3000TA-EU Electronics . 15
Figure 3-1: Front Panel of the Model 3000TA-EU ......................... 18
Figure 3-2: Required Front Door Clearance .................................. 18
Figure 3-3: Rear Panel of the Model 3000TA-EU .......................... 19
Figure 3-4: Equipment Interface Connector Pin Arrangement ....... 22
Figure 3-5: Remote Probe Connections ........................................ 27
Figure 3-6: FET Series Resistance ............................................... 27
Figure 5-1: Removing the Micro-Fuel ............................................ 66
Figure 5-2: Removing Fuse Block from Housing ........................... 69
Figure 5-3: Installing Fuses ........................................................... 70
Figure 5-4: Rear-Panel Screws ..................................................... 72
Figure A-1: Single and Dual 19" Rack Mounts .............................. 77
Teledyne Analytical Instruments
xi
3000TA- EU
List of Tables
Table 3-1: Analog Output Connections Pin Function .................... 23
Table 3-2: Alarm Relay Contact Pins ............................................ 24
Table 3-3: Remote Calibration Connections.................................. 25
Table 3-4: Range ID Relay Connections ....................................... 26
Table 3-5: Commands via RS-232 Input ....................................... 28
Table 5-1: Self Test Failure Codes................................................ 70
Teledyne Analytical Instruments
xii
Trace Oxygen Analyzer
Introduction
Introduction
1.1 Overview
The Teledyne Analytical Instruments Model 3000TA-EU Trace
Oxygen Analyzer is a versatile microprocessor-based instrument for
detecting oxygen at the parts-per-million (ppm) level in a variety of
gases. This manual covers the Model 3000TA-EU General Purpose
flush-panel and/or rack-mount units only. These units are for indoor use
in a nonhazardous environment.
1.2 Typical Applications
A few typical applications of the Model 3000TA-EU are:
• Monitoring inert gas blanketing
• Air separation and liquefaction
• Chemical reaction monitoring
• Semiconductor manufacturing
• Petrochemical process control
• Quality assurance
• Gas analysis certification.
1.3 Main Features of the Analyzer
The Model 3000TA-EU Trace Oxygen Analyzer is sophisticated
yet simple to use. The main features of the analyzer include:
• A 2-line alphanumeric display screen, driven by
microprocessor electronics, that continuously prompts
and informs the operator.
• High resolution, accurate readings of oxygen content
from low ppm levels through 25%. Large, bright, meter
readout.
• Nylon cell block. (Stainless steel optional)
Teledyne Analytical Instruments
1
Introduction
•
•
•
•
•
•
•
•
•
•
•
•
3000TA- EU
Advanced Micro-Fuel Cell, designed for trace analysis,
has a one year warranty and an expected lifetime of two
years.
Versatile analysis over a wide range of applications.
Microprocessor based electronics: 8-bit CMOS
microprocessor with 32 kB RAM and 128 kB ROM.
Three user definable output ranges (from 0-10 ppm
through 0- 250,000 ppm) allow best match to users
process and equipment.
Air-calibration range for convenient spanning at 20.9 %.
Auto Ranging allows analyzer to automatically select the
proper preset range for a given measurement. Manual
override allows the user to lock onto a specific range of
interest.
Two adjustable concentration alarms and a system failure
alarm.
Extensive self-diagnostic testing, at startup and on
demand, with continuous power-supply monitoring.
CE Compliance.
RS-232 serial digital port for use with a computer or
other digital communication device.
Four analog outputs: two for measurement (0–1 VDC and
Isolated 4–20 mA DC) and two for range identification.
Convenient and versatile, steel, flush-panel or rackmountable case with slide-out electronics drawer.
1.4 Model Designations
3000TA-EU: Standard model.
3000TA-EU-C: In addition to all standard features, this model
also has separate ports for zero and span gases,
and built-in control valves. The internal valves are
entirely under the control of the 3000TA-EU
electronics, to automatically switch between gases
in synchronization with the analyzer’s operations.
Teledyne Analytical Instruments
2
Trace Oxygen Analyzer
Introduction
1.5 Front Panel (Operator Interface)
The standard 3000TA-EU is housed in a rugged metal case with all
controls and displays accessible from the front panel. See Figure 1-1.
The front panel has thirteen buttons for operating the analyzer, a digital
meter, an alphanumeric display, and a window for viewing the sample
flowmeter.
Figure 1-1: Model 3000TA-EU Front Panel
Function Keys: Six touch-sensitive membrane switches are used to
change the specific function performed by the analyzer:
• Analyze Perform analysis for oxygen content of a sample
gas.
• System
Perform system-related tasks (described in detail
in chapter 4, Operation.).
• Span
Span calibrate the analyzer.
• Zero
Zero calibrate the analyzer.
• Alarms
Set the alarm setpoints and attributes.
• Range
Set up the 3 user definable ranges for the
instrument.
Data Entry Keys: Six touch-sensitive membrane switches are used to
input data to the instrument via the alphanumeric VFD display:
Teledyne Analytical Instruments
3
Introduction
3000TA- EU
•
Left & Right Arrows Select between functions currently
displayed on the VFD screen.
• Up & Down Arrows Increment or decrement values of
functions currently displayed.
• Enter Moves VFD display on to the next screen in a series.
If none remains, returns to the Analyze screen.
• Escape Moves VFD display back to the previous screen in a
series. If none remains, returns to the Analyze screen.
Digital Meter Display: The meter display is a LED device that
produces large, bright, 7-segment numbers that are legible in any
lighting. It produces a continuous readout from 0-10,000 ppm and then
switches to a continuous percent readout from 1-25%. It is accurate
across all analysis ranges without the discontinuity inherent in analog
range switching.
Alphanumeric Interface Screen: The VFD screen is an easy-to-use
interface from operator to analyzer. It displays values, options, and
messages that give the operator immediate feedback.
Flowmeter: Monitors the flow of gas past the sensor. Readout is 0.2 to
2.4 standard liters per minute (SLPM).
Standby Button: The Standby button turns off the display and
outputs, but circuitry is still operating.
CAUTION:
THE POWER CABLE MUST BE UNPLUGGED TO
FULLY DISCONNECT POWER FROM THE
INSTRUMENT. WHEN CHASSIS IS EXPOSED OR
WHEN ACCESS DOOR IS OPEN AND POWER
CABLE IS CONNECTED, USE EXTRA CARE TO
AVOID CONTACT WITH LIVE ELECTRICAL
CIRCUITS.
Access Door: For access to the Micro-Fuel Cell, the front panel swings
open when the latch in the upper right corner of the panel is pressed all
the way in with a narrow gauge tool. Accessing the main circuit board
requires unfastening rear panel screws and sliding the unit out of the
case.
1.6 Recognizing Difference Between LCD & VFD
LCD has GREEN background with BLACK characters. VFD has
DARK background with GREEN characters. In the case of VFD - NO
CONTRAST ADJUSTMENT IS NEEDED.
Teledyne Analytical Instruments
4
Trace Oxygen Analyzer
Introduction
1.7 Rear Panel (Equipment Interface)
The rear panel, shown in Figure 1-2, contains the gas and electrical
connectors for external inlets and outlets. Those that are optional are
shown shaded in the figure. The connectors are described briefly here
and in detail in the Installation chapter of this manual.
Figure 1-2: Model 3000 TA Rear Panel
•
•
•
•
•
Power Connection Universal AC power source.
Gas Inlet and Outlet One inlet (must be externally valved)
and one exhaust out. Three inlet when “C” option ordered.
RS-232 Port Serial digital concentration signal output and
control input.
Remote Valves Used in the 3000TA-EU for controlling
external solenoid valves only.
50-Pin Equipment Interface Port:
• Analog Outputs 0–1 VDC concentration plus 0-1 VDC
range ID, and isolated 4–20 mA DC plus 4-20 mA DC
range ID.
• Alarm Connections 2 concentration alarms and 1 system
alarm.
Teledyne Analytical Instruments
5
Introduction
•
•
•
•
•
3000TA- EU
Remote Span/Zero Digital inputs allow external control
of analyzer calibration.
Calibration Contact To notify external equipment that
instrument is being calibrated and readings are not
monitoring sample.
Range ID Contacts Four separate, dedicated, range relay
contacts. Low, Medium, High, Cal.
Network I/O Serial digital communications for local
network access. For future expansion. Not implemented
at this printing.
Optional:
• Calibration Gas Ports (Auto Cal Option) Separate
fittings for zero, span and sample gas input, and internal
valves for automatically switching the gases.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the
Model 3000TA-EU is accurate to 2-3 %. Accordingly,
internally scheduled calibrations can vary 2-3 % per day.
Teledyne Analytical Instruments
6
Trace Oxygen Analyzer
Operational Theory
Operational Theory
2.1 Introduction
The analyzer is composed of three subsystems:
1. Micro-Fuel Cell Sensor
2. Sample System
3. Electronic Signal Processing, Display and Control
The sample system is designed to accept the sample gas and
transport it through the analyzer without contaminating or altering the
sample prior to analysis. The Micro-Fuel Cell is an electrochemical
galvanic device that translates the amount of oxygen present in the
sample into an electrical current. The electronic signal processing,
display and control subsystem simplifies operation of the analyzer and
accurately processes the sampled data. The microprocessor controls all
signal processing, input/output and display functions for the analyzer.
2.2 Micro-Fuel Cell Sensor
2.2.1 Principles of Operation
The oxygen sensor used in the Model 3000T series is a Micro-Fuel
Cell designed and manufactured by Analytical Instruments. It is a sealed
plastic disposable electrochemical transducer.
The active components of the Micro-Fuel Cell are a cathode, an
anode, and the 15% aqueous KOH electrolyte in which they are
immersed. The cell converts the energy from a chemical reaction into an
electrical current in an external electrical circuit. Its action is similar to
that of a battery.
There is, however, an important difference in the operation of a
battery as compared to the Micro-Fuel Cell: In the battery, all reactants
are stored within the cell, whereas in the Micro-Fuel Cell, one of the
reactants (oxygen) comes from outside the device as a constituent of the
sample gas being analyzed. The Micro-Fuel Cell is therefore a hybrid
between a battery and a true fuel cell. (All of the reactants are stored
externally in a true fuel cell.)
Teledyne Analytical Instruments
7
Operational Theory
3000TA- EU
2.2.2 Anatomy of a Micro-Fuel Cell
The Micro-Fuel Cell is a cylinder only 11/4 inches in diameter and
11/4 inches thick. It is made of an extremely inert plastic, which can be
placed confidently in practically any environment or sample stream. It is
effectively sealed, although one end is permeable to oxygen in the
sample gas. The other end of the cell is a contact plate consisting of two
concentric foil rings. The rings mate with spring-loaded contacts in the
sensor block assembly and provide the electrical connection to the rest
of the analyzer. Figure 2-1 illustrates the external features.
Figure 2-1: Micro-Fuel Cell
Refer to Figure 2-2, Cross Section of a Micro-Fuel Cell, which
illustrates the following internal description.
Figure 2-2. Cross Section of a Micro-Fuel Cell (not to scale)
At the top end of the cell is a diffusion membrane of Teflon®,
whose thickness is very accurately controlled. Beneath the diffusion
membrane lies the oxygen sensing element—the cathode—with a
surface area almost 4 cm2. The cathode has many perforations to ensure
Teledyne Analytical Instruments
8
Trace Oxygen Analyzer
Operational Theory
sufficient wetting of the upper surface with electrolyte, and it is plated
with an inert metal.
The anode structure is below the cathode. It is made of lead and has
a proprietary design which is meant to maximize the amount of metal
available for chemical reaction.
At the rear of the cell, just below the anode structure, is a flexible
membrane designed to accommodate the internal volume changes that
occur throughout the life of the cell. This flexibility assures that the
sensing membrane remains in its proper position, keeping the electrical
output constant.
The entire space between the diffusion membrane, above the
cathode, and the flexible rear membrane, beneath the anode, is filled
with electrolyte. Cathode and anode are submerged in this common
pool. They each have a conductor connecting them to one of the external
contact rings on the contact plate, which is on the bottom of the cell.
2.2.3 Electrochemical Reactions
The sample gas diffuses through the Teflon membrane. Any
oxygen in the sample gas is reduced on the surface of the cathode by the
following HALF REACTION:
O2 + 2H2O + 4e– → 4OH–
(cathode)
(Four electrons combine with one oxygen molecule—in the
presence of water from the electrolyte—to produce four hydroxyl ions.)
When the oxygen is reduced at the cathode, lead is simultaneously
oxidized at the anode by the following HALF REACTION:
Pb + 2OH– → Pb+2 + H2O + 2e–
(anode)
(Two electrons are transferred for each atom of lead that is
oxidized. Therefore it takes two of the above anode reactions to balance
one cathode reaction and transfer four electrons.)
The electrons released at the surface of the anode flow to the cathode
surface when an external electrical path is provided. The current is
proportional to the amount of oxygen reaching the cathode. It is measured
and used to determine the oxygen concentration in the gas mixture.
Teledyne Analytical Instruments
9
Operational Theory
3000TA- EU
The overall reaction for the fuel cell is the SUM of the half
reactions above, or:
2Pb + O2 →2PbO
(These reactions will hold as long as no gaseous components
capable of oxidizing lead—such as iodine, bromine, chlorine and
fluorine—are present in the sample.)
The output of the fuel cell is limited by (1) the amount of oxygen in
the cell at the time and (2) the amount of stored anode material.
In the absence of oxygen, no current is generated.
2.2.4 The Effect of Pressure
In order to state the amount of oxygen present in the sample in
parts-per-million or a percentage of the gas mixture, it is necessary that
the sample diffuse into the cell under constant pressure.
If the total pressure increases, the rate that oxygen reaches the
cathode through the diffusing membrane will also increase. The electron
transfer, and therefore the external current, will increase, even though
the oxygen concentration of the sample has not changed. It is therefore
important that the sample pressure at the fuel cell (usually vent pressure)
remain relatively constant between calibrations.
2.2.5 Calibration Characteristics
Given that the total pressure of the sample gas on the surface of the
Micro-Fuel Cell input is constant, a convenient characteristic of the cell is
that the current produced in an external circuit is directly proportional to the
rate at which oxygen molecules reach the cathode, and this rate is directly
proportional to the concentration of oxygen in the gaseous mixture. In other
words it has a linear characteristic curve, as shown in Figure 2-3. Measuring
circuits do not have to compensate for nonlinearities.
In addition, since there is zero output in the absence oxygen, the
characteristic curve has close to an absolute zero (within ± 1 ppm
oxygen). In practical application, zeroing may still used to compensate
for the combined zero offsets of the cell and the electronics. (The
electronics is zeroed automatically when the instrument power is turned
on.)
Teledyne Analytical Instruments
10
Trace Oxygen Analyzer
Operational Theory
Figure 2-3. Characteristic Input/Output Curve for a Micro-Fuel
Cell
2.3 Sample System
The sample system delivers gases to the Micro-Fuel Cell sensor
from the analyzer rear panel inlet. Depending on the mode of operation
either sample or calibration gas is delivered.
The Model 3000TA-EU sample system is designed and fabricated
to ensure that the oxygen concentration of the gas is not altered as it
travels through the sample system. The sample encounters almost no
dead space. This minimizes residual gas pockets that can interfere with
trace analysis.
The sample system for the standard instrument incorporates 1/4
inch tube fittings for sample inlet and outlet connections at the rear
panel. For metric system installations, 6 mm adapters are supplied with
each instrument to be used if needed. The sample or calibration gas
flows through the system is monitored by a flowmeter downstream from
the cell. Figure 2-4 shows the piping layout and flow diagram for the
standard model.
Teledyne Analytical Instruments
11
Operational Theory
3000TA- EU
Figure 2-4: Piping Layout and Flow Diagram for Standard Model
Figure 2-5 is the flow diagram for the sampling system. In the standard instrument, calibration gases (zero and span) can be connected directly to the Sample In port by teeing to the port with appropriate valves.
The shaded portion of the diagram shows the components added when
the –C option is ordered. The valving is installed inside the 3000TAEU-C enclosure and is regulated by the instruments internal electronics.
Teledyne Analytical Instruments
12
Trace Oxygen Analyzer
Operational Theory
Figure 2-5: Flow Diagram
2.4 Electronics and Signal Processing
The Model 3000TA-EU Trace Oxygen Analyzer uses an 8031
microcontroller with 32 kB of RAM and 128 kB of ROM to control all
signal processing, input/output, and display functions for the analyzer.
System power is supplied from a universal power supply module
designed to be compatible with any international power source. Figure
2-6 shows the location of the power supply and the main electronic PC
boards.
The signal processing electronics including the microprocessor,
analog to digital, and digital to analog converters are located on the
motherboard at the bottom of the case. The preamplifier board is
mounted on top of the motherboard as shown in the figure. These boards
are accessible after removing the back panel. Figure 2-7 is a block
diagram of the Analyzer electronics.
Teledyne Analytical Instruments
13
Operational Theory
3000TA- EU
Figure 2-6: 3000TA-EU Internal Electronic Component Location
In the presence of oxygen the cell generates a current. A current to
voltage amplifier converts this current to a voltage, which is amplified in
the second stage amplifier.
The second stage amplifier also supplies temperature compensation
for the oxygen sensor output. This amplifier circuit incorporates a thermistor, which is physically located in the cell block. The thermistor is a
temperature dependent resistance that changes the gain of the amplifier
in proportion to the temperature changes in the block. This change is inversely proportional to the change in the cell output due to the same
temperature changes. The result is a signal that is temperature
independent. The output from the second stage amplifier is sent to an 18
bit analog to digital converter controlled by the microprocessor.
Teledyne Analytical Instruments
14
Trace Oxygen Analyzer
Operational Theory
Figure 2-7: Block Diagram of the Model 3000TA-EU Electronics
Teledyne Analytical Instruments
15
Operational Theory
3000TA- EU
The digital concentration signal along with input from the control
panel is processed by the microprocessor, and appropriate control
signals are directed to the display, alarms and communications port. The
same digital information is also sent to a 12 bit digital to analog
converter that produces the 4-20 mA DC and the 0-1 VDC analog
concentration signal outputs, and the analog range ID outputs.
Signals from the power supply are also monitored, and through the
microprocessor, the system failure alarm is activated if a malfunction is
detected.
Teledyne Analytical Instruments
16
Trace Oxygen Analyzer
Installation
Installation
Installation of the Model 3000TA-EU Analyzer includes:
1. Unpacking
2. Mounting
3. Gas connections
4. Electrical connections
5. Installing the Micro-Fuel Cell
6. Testing the system.
3.1 Unpacking the Analyzer
Although the analyzer is shipped complete, certain parts, such as
fuses and sensors, are wrapped separately to be installed on site as part
of the installation. Carefully unpack the analyzer and inspect it for
damage. Immediately report any damage or shortages to the shipping
agent.
3.2 Mounting the Analyzer
The Model 3000TA-EU is for indoor use in a general purpose area.
It is NOT for hazardous environments of any type.
The standard model is designed for flush panel mounting. Figure 31 is an illustration of the 3000TA-EU standard front panel and mounting
bezel. There are four mounting holes—one in each corner of the rigid
frame. The Drawings section in the rear of this manual contains outline
dimensions and mounting hole spacing diagrams.
On special order, a 19" rack-mounting panel can be provided. For
rack mounting, one or two 3000 series analyzers are flush-panel
mounted on the rack panel. See Appendix for dimensions of the
mounting panel.
Teledyne Analytical Instruments
17
Installation
3000TA- EU
Figure 3-1: Front Panel of the Model 3000TA-EU
All operator controls are mounted on the control panel, which is
hinged on the left edge and doubles as the door that provides access to
the sensor and cell block inside the instrument. The door is spring
loaded and will swing open when the button in the center of the latch
(upper right corner) is pressed all the way in with a narrow gauge tool
(less than 0.18 inch wide), such as a small hex wrench or screwdriver
Allow clearance for the door to open in a 90-degree arc of radius 7.125
inches. See Figure 3-2.
Figure 3-2: Required Front Door Clearance
Teledyne Analytical Instruments
18
Trace Oxygen Analyzer
Installation
3.3 Rear Panel Connections
Figure 3-3 shows the Model 3000TA-EU rear panel. There are
ports for gas, power, and equipment interface. The Zero In and Span In
ports are not included on the standard model, but are available as
options.
Figure 3-3: Rear Panel of the Model 3000TA-EU
3.3.1 Gas Connections
Before using this instrument, it should be determined if the unit will
be used for pressurized service or vacuum service and low pressure
applications. Inspect the restrictor kit that came with the unit. The kit
consists of two restrictors and a union for 1/4” diameter tubing. Notice
that the two 1-3/4” long, 1/4” diameter tubing are restrictors. It has an
open end and a closed end with a small circular orifice. The restrictor
without the blue sticker is for low pressure and vacuum service. For high
pressure applications (5-50psig), use the restrictor that has a blue sticker
on the body.
Teledyne Analytical Instruments
19
Installation
3000TA- EU
For pressurized service (> 5psig) , use the restrictor with the blue
dot and union from the restrictor kit and attach it to the Sample In port.
The small circular orifice should face away from the back of the unit
(against the direction of gas flow).
For positive pressures less than 5 psig use the low-pressure
restrictor without the blue dot in the Sample-in line.
For vacuum service (5-10 in Hg), use the restrictor without the
blue dot sticker and union but attach it to the Exhaust Out port. The
small circular orifice should face toward the back of the unit (against the
direction of gas flow).
Remove the blue sticker from the restrictor before using.
Warning:
Operating the unit without restrictors can cause
damage to the micro-fuel cell.
The unit is manufactured with 1/4 inch tube fittings, and 6 mm
adapters are supplied for metric system installations. For a safe
connection:
1. Insert the tube into the tube fitting, and finger-tighten the nut
until the tubing cannot be rotated freely, by hand, in the
fitting. (This may require an additional 1/8 turn beyond
finger-tight.)
2. Hold the fitting body steady with a backup wrench, and with
another wrench rotate the nut another 1-1/4 turns.
SAMPLE IN:
In the standard model, gas connections are made at the SAMPLE IN
and EXHAUST OUT connections. Calibration gases must be tee'd into
the Sample inlet with appropriate valves.
The gas pressure in should be reasonably regulated. Pressures
between 2 and 50 psig are acceptable as long as the pressure, once
established, will keep the front panel flowmeter reading in an acceptable
range (0.1 to 2.4 SLPM). For non-pressurized sample or very low
pressure, (2 psig or less) vacuum service plumbing is recommended.
Exact figures will depend on your process.
If greater flow is required for improved response time, install a
bypass in the sampling system upstream of the analyzer input.
Teledyne Analytical Instruments
20
Trace Oxygen Analyzer
Installation
EXHAUST OUT:
Exhaust connections must be consistent with the hazard level of the
constituent gases. Check Local, State, and Federal laws, and ensure that
the exhaust stream vents to an appropriately controlled area if required.
ZERO IN and SPAN IN (Optional):
These are additional ports for inputting span gas and zero gas.
There are electrically operated valves inside for automatic switching
between sample and calibration gases. These valves are completely
under control of the 3000T Electronics. They can be externally
controlled only indirectly through the Remote Cal Inputs, described
below.
Pressure, flow, and safety considerations are the same as prescribed
for the SAMPLE IN inlet, above.
3.3.2 Electrical Connections
For safe connections, no uninsulated wiring should be able to come
in contact with fingers, tools or clothing during normal operation.
CAUTION:
USE SHIELDED CABLES. ALSO, USE PLUGS THAT
PROVIDE EXCELLENT EMI/RFI PROTECTION. THE
PLUG CASE MUST BE CONNECTED TO THE CABLE
SHIELD, AND IT MUST BE TIGHTLY FASTENED TO
THE ANALYZER WITH ITS FASTENING SCREWS.
ULTIMATELY, IT IS THE INSTALLER WHO ENSURES
THAT THE CONNECTIONS PROVIDE ADEQUATE
EMI/RFI SHIELDING.
3.3.2.1 PRIMARY INPUT POWER
The power cord receptacle and fuse block are located in the same
assembly. Insert the power cord into the power cord receptacle.
CAUTION:
POWER IS APPLIED TO THE INSTRUMENT'S
CIRCUITRY AS LONG AS THE INSTRUMENT IS
CONNECTED TO THE POWER SOURCE. THE RED
SWITCH ON THE FRONT PANEL IS FOR SWITCHING
POWER ON OR OFF TO THE DISPLAYS AND OUTPUTS ONLY.
The universal power supply requires a 85–250 V ac, 47-63 Hz
power source.
Teledyne Analytical Instruments
21
Installation
3000TA- EU
Fuse Installation: The fuse block, at the right of the power cord
receptacle, accepts US or European size fuses. A jumper replaces the
fuse in whichever fuse receptacle is not used. Fuses are not installed at
the factory. Be sure to install the proper fuse as part of installation. (See
Fuse Replacement in chapter 5, maintenance.)
3.3.2.2 50-PIN EQUIPMENT INTERFACE CONNECTOR
Figure 3-4 shows the pin layout of the Equipment Interface
connector. The arrangement is shown as seen when the viewer faces the
rear panel of the analyzer. The pin numbers for each input/output
function are given where each function is described in the paragraphs
below.
Figure 3-4: Equipment Interface Connector Pin Arrangement
Analog Outputs: There are four DC output signal pins—two pins
per output. For polarity, see Table 3-1. The outputs are:
0–1 VDC % of Range: Voltage rises linearly with increasing
oxygen, from 0 V at 0 ppm to 1 V at full
scale ppm. (Full scale = 100% of
programmable range.)
0–1 VDC Range ID:
0.25 V = Low Range, 0.5 V = Medium
Range, 0.75 V = High Range, 1 V = Air
Cal Range.
4–20 mA DC % Range: Current increases linearly with
increasing oxygen, from 4 mA at 0 ppm
to 20 mA at full scale ppm. (Full scale =
100% of programmable range.)
4–20 mA DC Range ID: 8 mA = Low Range, 12 mA = Medium
Range, 16 mA = High Range, 20 mA =
Air Cal Range.
Teledyne Analytical Instruments
22
Trace Oxygen Analyzer
Installation
Table 3-1: Analog Output Connections Pin Function
Pin
Function
3
+ Range ID, 4-20 mA, floating
4
– Range ID, 4-20 mA, floating
5
+ % Range, 4-20 mA, floating
6
– % Range, 4-20 mA, floating
8
+ Range ID, 0-1 VDC
23
– Range ID, 0-1 VDC, negative ground
24
7
+ % Range, 0-1 VDC
– % Range, 0-1 VDC, negative ground
Alarm Relays: The nine alarm-circuit connector pins connect to
the internal alarm relay contacts. Each set of three pins provides one set
of Form C relay contacts. Each relay has both normally open and
normally closed contact connections. The contact connections are shown
in Table 3-2. They are capable of switching up to 3 amperes at 250 V ac
into a resistive load. The connectors are:
• Threshold Alarm 1:
• Can be configured as high (actuates when concentration
is above threshold), or low (actuates when concentration
is below threshold).
• Can be configured as failsafe or non-failsafe.
• Can be configured as latching or non-latching.
• Can be configured out (defeated).
• Threshold Alarm 2:
• Can be configured as high (actuates when concentration
is above threshold), or low (actuates when concentration
is below threshold).
• Can be configured as failsafe or non-failsafe.
• Can be configured as latching or non-latching.
• Can be configured out (defeated).
• System Alarm:
• Actuates when DC power supplied to circuits is
unacceptable in one or more parameters. Permanently
Teledyne Analytical Instruments
23
Installation
•
3000TA- EU
configured as failsafe and latching. Cannot be defeated.
Actuates if self test fails.
Reset by pressing
button to remove power. Then press
again and any other button EXCEPT System to
resume.
Further detail can be found in Chapter 4, Section 4-5.
Table 3-2: Alarm Relay Contact Pins
Pin
Contact
45
28
Threshold Alarm 1, normally closed contact
Threshold Alarm 1, moving contact
46
Threshold Alarm 1, normally open contact
42
44
Threshold Alarm 2, normally closed contact
Threshold Alarm 2, moving contact
43
Threshold Alarm 2, normally open contact
36
System Alarm, normally closed contact
20
37
System Alarm, moving contact
System Alarm, normally open contact
Digital Remote Cal Inputs: Accept 0 V (off) or 24 VDC (on)
inputs for remote control of calibration. (See Remote Calibration
Protocol below.) See Table 3-3 for pin connections.
Zero: Floating input. 5 to 24 V input across the + and – pins puts the
analyzer into the Zero mode. Either side may be grounded at
the source of the signal. 0 to 1 volt across the terminals allows
Zero mode to terminate when done. A synchronous signal must
open and close the external zero valve appropriately. See
Remote Probe Connector. (The –C option internal valves
operate automatically.)
Span:
Floating input. 5 to 24 V input across the + and – pins puts the
analyzer into the Span mode. Either side may be grounded at the
source of the signal. 0 to 1 volt across the terminals allows Span
mode to terminate when done. A synchronous signal must open
and close external span valve appropriately. See Figure 3-5
Teledyne Analytical Instruments
24
Trace Oxygen Analyzer
Installation
Remote Probe Connector. (The –C option internal valves
operate automatically.)
Cal Contact: This relay contact is closed while analyzer is
spanning and/or zeroing. (See Remote Calibration Protocol below.)
Table 3-3: Remote Calibration Connections
Pin
9
11
10
12
40
41
Function
+ Remote Zero
– Remote Zero
+ Remote Span
– Remote Span
Cal Contact
Cal Contact
Remote Calibration Protocol: To properly time the Digital
Remote Cal Inputs to the Model 3000TA-EU Analyzer, the customer's
controller must monitor the Cal Relay Contact.
When the contact is OPEN, the analyzer is analyzing, the Remote
Cal Inputs are being polled, and a zero or span command can be sent.
When the contact is CLOSED, the analyzer is already calibrating. It
will ignore your request to calibrate, and it will not remember that
request.
Once a zero or span command is sent, and acknowledged (contact
closes), release it. If the command is continued until after the zero or
span is complete, the calibration will repeat and the Cal Relay Contact
(CRC) will close again.
For example:
1. Test the CRC. When the CRC is open, Send a zero command
until the CRC closes (The CRC will quickly close.)
2. When the CRC closes, remove the zero command.
3. When CRC opens again, send a span command until the
CRC closes. (The CRC will quickly close.)
4. When the CRC closes, remove the span command.
When CRC opens again, zero and span are
done, and the sample is being analyzed.
Teledyne Analytical Instruments
25
Installation
3000TA- EU
Note: The Remote Valve connections (described below) provides
signals to ensure that the zero and span gas valves will be
controlled synchronously. If you have the –C Internal valve
option—which includes additional zero and span gas
inputs— the 3000T automatically regulates the zero, span
and sample gas flow.
Range ID Relays: Four dedicated Range ID relay contacts. The
first three ranges are assigned to relays in ascending order—Low range
is assigned to Range 1 ID, Medium range is assigned to Range 2 ID, and
High range is assigned to Range 3 ID. The fourth range is reserved for
the Air Cal Range (25%). Table 3-4 lists the pin connections.
Table 3-4: Range ID Relay Connections
Pin
Function
21
Range 1 ID Contact
38
22
Range 1 ID Contact
Range 2 ID Contact
39
Range 2 ID Contact
19
Range 3 ID Contact
18
34
Range 3 ID Contact
Range 4 ID Contact (Air Cal)
35
Range 4 ID Contact (Air Cal)
Network I/O: A serial digital input/output for local network
protocol. At this printing, this port is not yet functional. It is to be used
for future options to the instrument. Pins 13 (+) and 29 (–).
Remote Valve Connections: The 3000TA-EU is a single-chassis
instrument, which has no Remote Valve Unit. Instead, the Remote Valve
connections are used as a method for directly controlling external
sample/zero/span gas valves. See Figure 3-5.
Teledyne Analytical Instruments
26
Trace Oxygen Analyzer
Installation
Figure 3-5: Remote Probe Connections
The voltage from these outputs is nominally 0 V for the OFF and
15 VDC for the ON conditions. The maximum combined current that
can be pulled from these output lines is 100 mA. (If two lines are ON at
the same time, each must be limited to 50 mA, etc.) If more current
and/or a different voltage is required, use a relay, power amplifier, or
other matching circuitry to provide the actual driving current.
In addition, each individual line has a series FET with a nominal
ON resistance of 5 ohms (9 ohms worst case). This can limit the
obtainable voltage, depending on the load impedance applied. See
Figure 3-6.
Figure 3-6: FET Series Resistance
3.3.2.3 RS-232 PORT
The digital signal output is a standard, full duplex RS-232 serial
communications port used to connect the analyzer to a computer,
Teledyne Analytical Instruments
27
Installation
3000TA- EU
terminal, or other digital device. It requires a standard 9-pin D
connector.
The output data is status information, in digital form, updated every
two seconds. Status is reported in the following order:
• The concentration in ppm or percent
• The range in use (HI, MED, LO)
• The span of the range (0-100 ppm, etc)
• Which alarms—if any—are disabled (AL–x DISABLED)
• Which alarms—if any—are tripped (AL–x ON).
Each status output is followed by a carriage return and line feed.
Three input functions using RS-232 have been implemented to
date. They are described in Table 3-5.
Table 3-5: Commands via RS-232 Input
Command
as
az
Description
Immediately starts an autospan.
Immediately starts an autozero.
st
Toggling input. Stops/Starts any status
message output from the RS-232, until
st is sent again.
The RS-232 protocol allows some flexibility in its implementation.
Table 3-6 lists certain RS-232 values that are required by the 3000TAEU implementation.
Table 3-6: Required RS-232 Options
Parameter
Setting
Baud
2400
Byte
8 bits
Parity
none
Stop Bits
Message Interval
1
2 seconds. When CRC opens
again, zero and span are done,
Teledyne Analytical Instruments
28
Trace Oxygen Analyzer
Installation
3.4 Installing the Micro-Fuel Cell
The Micro-Fuel Cell is not installed in the cell block when the
instrument is shipped. Install it before the analyzer is placed in service.
Refer to the procedure described in Section 5.2. Note that there are
different installations procedures depending on the type of cell your
application requires. See Section 5.2 for more information.
Once it is expended, or if the cell is exposed to air for too long, the
Micro-Fuel Cell will need to be replaced. The cell could also require
replacement if the instrument has been idle for too long.
When the Micro-Fuel Cell needs to be installed or replaced, follow
the procedures in Chapter 5, Maintenance, for removing and installing
cells.
3.5 Testing the System
Before plugging the instrument into the power source:
• Check the integrity and accuracy of the gas connections.
Make sure there are no leaks.
• Check the integrity and accuracy of the electrical
connections. Make sure there are no exposed conductors
• Verify that the restriction device has been properly installed
(see section 3.3.1).
• Check that inlet sample pressure is within the accepted range
(se section 3.3.1).
• Power up the system, and test it by repeating the SelfDiagnostic Test as described in Chapter 4, Section 4.3.5.
Teledyne Analytical Instruments
29
Trace Oxygen Analyzer
Operation
Operation
4.1 Introduction
Once the analyzer has been installed, it can be configured for your
application. To do this you will:
• Set system parameters:
• Establish a security password, if desired, requiring
Operator to log in.
• Establish and start an automatic calibration cycle, if
desired.
• Calibrate the instrument.
• Define the three user selectable analysis ranges. Then choose
autoranging or select a fixed range of analysis, as required.
• Set alarm setpoints, and modes of alarm operation (latching,
failsafe, etc).
Before you configure your 3000TA-EU these default values are in
effect:
Ranges: LO = 100 ppm, MED = 1000 ppm, HI = 10,000 ppm.
Auto Ranging: ON
Alarm Relays: Defeated, 1000 ppm, HI, Not failsafe, Not
latching.
Zero: Auto, every 0 days at 0 hours.
Span: Auto, at 000008.00 ppm, every 0 days at 0 hours.
If you choose not to use password protection, the default password
is automatically displayed on the password screen when you start up,
and you simply press Enter for access to all functions of the analyzer.
4.2 Using the Data Entry and Function Buttons
Data Entry Buttons: The ◄►arrow buttons select options from
the menu currently being displayed on the VFD screen. The selected
option blinks.
Teledyne Analytical Instruments
31
Operation
3000TA- EU
When the selected option includes a modifiable item, the ▲/▼
arrow buttons can be used to increment or decrement that modifiable
item.
The Enter button is used to accept any new entries on the VFD
screen. The Escape button is used to abort any new entries on the VFD
screen that are not yet accepted by use of the Enter button.
Figure 4-1 shows the hierarchy of functions available to the
operator via the function buttons. The six function buttons on the
analyzer are:
• Analyze. This is the normal operating mode. The analyzer
monitors the oxygen content of the sample, displays the
percent of oxygen, and warns of any alarm conditions.
• System. The system function consists of six subfunctions
that regulate the internal operations of the analyzer:
• Auto-Cal setup
• Password assignment
• Self-Test initiation
• Checking software version
• Logging out.
• Zero. Used to set up a zero calibration.
• Span. Used to set up a span calibration.
• Alarms. Used to set the alarm setpoints and determine
whether each alarm will be active or defeated, HI or LO
acting, latching, and/or failsafe.
• Range. Used to set up three analysis ranges that can be
switched automatically with auto-ranging or used as
individual fixed ranges.
Any function can be selected at any time by pressing the
appropriate button (unless password restrictions apply). The order as
presented in this manual is appropriate for an initial setup.
Each of these functions is described in greater detail in the
following procedures. The VFD screen text that accompanies each
operation is reproduced, at the appropriate point in the procedure, in a
Monospaced type style. Pushbutton names are printed in Oblique type.
Teledyne Analytical Instruments
32
Trace Oxygen Analyzer
Operation
Figure 4-1: Hierarchy of Functions and Sub functions
4.3 The System Function
The subfuctions of the System function are described below.
Specific procedures for their use follow the descriptions:
• Auto-Cal: Used to define an automatic calibration sequence
and/or start an Auto-Cal.
• PSWD: Security can be established by choosing a 5 digit
password (PSWD) from the standard ASCII character set.
Teledyne Analytical Instruments
33
Operation
3000TA- EU
(See Installing or Changing a Password, below, for a table of
ASCII characters available.) Once a unique password is
assigned and activated, the operator MUST enter the
UNIQUE password to gain access to set-up functions which
alter the instrument's operation, such as setting the instrument
span or zero setting, adjusting the alarm setpoints, or defining
analysis ranges.
After a password is assigned, the operator must log out to
activate it. Until then, anyone can continue to operate the
instrument without entering the new password.
•
•
•
•
•
•
Only one password can be defined. Before a unique
password is assigned, the system assigns TETAI by default.
This allows access to anyone. After a unique password is
assigned, to defeat the security, the password must be
changed back to TETAI.
Logout: Logging out prevents an unauthorized tampering
with analyzer settings.
More: Select and enter More to get a new screen with
additional subfunctions listed.
Self–Test: The instrument performs a self-diagnostic test to
check the integrity of the power supply, output boards and
amplifiers.
Version: Displays Manufacturer, Model, and Software
Version of the instrument.
Show Negative: The operator selects whether display can
show negative oxygen readings or not.
TRAK/HLD: The operator sets whether the instrument
analog outputs track the concentration change during
calibration and sets a time delay for the concentration alarms
after calibration
4.3.1 Tracking Oxygen Readings During Calibration and
Alarm Delay
The user has the option of setting the preference as to whether the
analog outputs track the display readings during calibration or not. To
set the preference, press the System key once and the first System menu
will appear in the VFD display:
Teledyne Analytical Instruments
34
Trace Oxygen Analyzer
Operation
TRAK/HLD Auto-Cal
PSWD Logout More
TRAK/HLD should be blinking. To enter this system menu press
the Enter key once:
Output Sttng: TRACK
Alarm Dly: 10 min
or
Output Sttng: HOLD
Alarm Dly: 10 min
In the first line, TRACK or HOLD should be blinking. The
operator can toggle between TRACK and HOLD with the Up or Down
keys. When TRACK is selected, the analog outputs (0-1 VDC and 4-20
mA) and the range ID contacts will track the instrument readings during
calibration (either zero or span). TRACK is the factory default.
When HOLD is selected, the analog outputs (0-1 VDC and 4-20
mA) and the range ID contacts will freeze on their last state before
entering one of the calibration modes. When the instrument returns to
the Analyze mode, either by a successful or an aborted calibration, there
will be a three-minute delay before the analog outputs and the range ID
contacts start tracking again.
The concentration alarms freeze on their last state before entering
calibration regardless of selecting HOLD or TRACK. But, when HOLD
is selected the concentration alarms will remain frozen for the time
displayed in the second line of the TRAK/HLD menu after the analyzer
returns to the Analyze mode.
The factory default is three minutes, but the delay time is programmable. To adjust to delay time use the Left or Right arrow keys. When
the time displayed on the second line blinks, it can be adjusted by
Pressing the Up or Down keys to increase or decrease its value. The
minimum delay is 1 minute, the maximum is 30.
This preference is stored in non-volatile memory so that it is
recovered if power is removed from the instrument.
Teledyne Analytical Instruments
35
Operation
3000TA- EU
4.3.2 Setting up an Auto-Cal
When proper automatic valving is connected (see Chapter 3,
Installation), the analyzer can cycle itself through a sequence of steps
that automatically zero and span the instrument.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the
Model 3000TA-EU is accurate to 2-3 %. Accordingly,
internally scheduled calibrations can vary 2-3 % per day.
To setup an Auto–Cal cycle:
Choose System from the Function buttons. The LCD will display
five subfunctions.
TRAK/HLD Auto—Cal
PSWD Logout More
Use ◄►arrows to blink Auto—Cal, and press Enter. A new screen
for Span/Zero set appears.
Span OFF Nxt: 0d 0h
Zero OFF Nxt: 0d 0h
Press ◄►arrows to blink Span (or Zero), then press Enter again.
(You won’t be able to set OFF to ON if a zero interval is entered.) A
Span Every ... (or Zero Every ...) screen appears.
Span Every 0 d
Start 0 h from now
Use ▲/▼ arrows to set an interval value, then use ◄► arrows to
move to the start-time value. Use ▲/▼ arrows to set a start-time value.
To turn ON the Span and/or Zero cycles (to activate Auto-Cal):
Press System again, choose Auto—Cal, and press Enter again. When
the Span/ Zero values screen appears, use the ◄► arrows to blink the
Span (or Zero) OFF/ON field. Use ▲/▼ arrows to set the OFF/ON
field to ON. You can now turn these fields ON because there is a
nonzero span interval defined.
Teledyne Analytical Instruments
36
Trace Oxygen Analyzer
Operation
4.3.3 Password Protection
If a password is assigned, then setting the following system
parameters can be done only after the password is entered: span and
zero settings, alarm setpoints, analysis range definitions, switching
between autoranging and manual override, setting up an auto-cal, and
assigning a new password. However, the instrument can still be used for
analysis or for initiating a self- test without entering the password.
If you have decided not to employ password security, use the
default password TETAI. This password will be displayed automatically
by the microprocessor. The operator just presses the Enter key to be
allowed total access to the instrument’s features.
Note: If you use password security, it is advisable to keep a copy
of the password in a separate, safe location.
4.3.3.1 ENTERING THE PASSWORD
To install a new password or change a previously installed
password, you must key in and ENTER the old password first. If the
default password is in effect, pressing the ENTER button will enter the
default TETAI password for you.
Press System to enter the System mode.
TRAK/HLD Auto—Cal
PSWD Logout More
Use the ◄►arrow keys to scroll the blinking over to PSWD, and
press Enter to select the password function. Either the default TETAI
password or AAAAA place holders for an existing password will appear
on screen depending on whether or not a password has been previously
installed.
TETAI
Enter PWD
or
Teledyne Analytical Instruments
37
Operation
3000TA- EU
AAAAA
Enter PWD
The screen prompts you to enter the current
password. If you are not using password protection,
press Enter to accept TETAI as the default password.
If a password has been previously installed, enter the
password using the ◄►arrow keys to scroll back and
forth between letters, and the ▲/▼ arrow keys to
change the letters to the proper password. Press Enter
to enter the password.
If the password is accepted, the screen will
indicate that the password restrictions have been
removed and you have clearance to proceed.
PSWD Restrictions
Removed
In a few seconds, you will be given the opportunity to change this
password or keep it and go on.
Change Password?
=Yes
=No
Press Escape to move on, or proceed as in Changing the
Password, below.
4.3.3.2 INSTALLING OR CHANGING THE PASSWORD
If you want to install a password, or change an existing password,
proceed as above in Entering the Password. When you are given the
opportunity to change the password:
Change Password?
=Yes
=No
Press Enter to change the password (either the default TETAI or
the previously assigned password), or press Escape to keep the existing
password and move on.
If you chose Enter to change the password, the password
assignment screen appears.
Teledyne Analytical Instruments
38
Trace Oxygen Analyzer
Operation
TETAI
To Proceed
or
AAAAA
To Proceed
Enter the password using the ◄►arrow keys to move back and
forth between the existing password letters, and the ▲/▼ arrow keys to
change the letters to the new password. The full set of 94 characters
available for password use are shown in the table below.
Characters Available for Password Definition:
A
K
U
i
s
}
)
3
=
B
L
V
`
j
t
→
*
4
>
C
M
W
a
k
u
!
+
5
?
D
N
X
b
l
v
"
'
6
@
E
O
Y
c
m
w
#
7
F
P
Z
d
n
x
$
.
8
G
Q
[
e
o
y
%
/
9
H
R
¥
f
p
z
&
0
:
I
S
]
g
q
{
'
1
;
J
T
^
h
r
|
(
2
<
When you have finished typing the new password, press Enter. A
verification screen appears. The screen will prompt you to retype your
password for verification.
AAAAA
Retype PWD To Verify
Wait a moment for the entry screen. You will be given clearance to
proceed.
AAAAA
TO Proceed
Teledyne Analytical Instruments
39
Operation
3000TA- EU
Use the arrow keys to retype your password and press Enter when
finished. Your password will be stored in the microprocessor and the
system will immediately switch to the Analyze screen, and you now
have access to all instrument functions.
If all alarms are defeated, the Analyze screen appears as:
0.0
ppm
Anlz
Range: 0 — 100
If an alarm is tripped, the second line will change to show which
alarm it is:
0.0
ppm
Anlz
AL—1
Note: If you log off the system using the logout function in the
system menu, you will now be required to re-enter the
password to gain access to Span, Zero, Alarm, and Range
functions.
4.3.4 Logout
The Logout function provides a convenient means of leaving the
analyzer in a password protected mode without having to shut the
instrument off. By entering Logout, you effectively log off the
instrument leaving the system protected against use until the password is
reentered. To log out, press the System button to enter the System
function.
TRAK/HLD Auto—Cal
PSWD Logout More
Use the ◄►arrow keys to position the blinking over the Logo ut
function, and press Enter to Log out. The screen will display the
message:
Teledyne Analytical Instruments
40
Trace Oxygen Analyzer
Operation
Protected Until
Password Reentered
4.3.5 System Self-Diagnostic Test
The Model 3000TA-EU has a built-in self-diagnostic testing
routine. Pre-programmed signals are sent through the power supply,
output board and sensor circuit. The return signal is analyzed, and at the
end of the test the status of each function is displayed on the screen,
either as OK or as a number between 1 and 3. (See System Self
Diagnostic Test in Chapter 5 for number code).
The self diagnostics are run automatically by the analyzer
whenever the instrument is turned on, but the test can also be run by the
operator at will. To initiate a self diagnostic test during operation:
Press the System button to start the System function.
TRAK/HLD Auto—Cal
PSWD Logout More
Use the ◄►arrow keys to blink More, then press Enter.
Version Self—Test
Use the ◄►arrow keys again to move the blinking to the Self–
Test function. The screen will follow the running of the diagnostic.
RUNNING DIAGNOSTIC
Testing Preamp — 83
During preamp testing there is a countdown in the lower right
corner of the screen. When the testing is complete, the results are
displayed.
Power: OK
Analog: OK
Preamp: 3
Teledyne Analytical Instruments
41
Operation
3000TA- EU
The module is functioning properly if it is followed by OK. A
number indicates a problem in a specific area of the instrument. Refer to
Chapter 5 Maintenance and Troubleshooting for number-code
information. The results screen alternates for a time with:
Press Any Key
To Continue...
Then the analyzer returns to the initial System screen.
4.3.6 Version Screen
Move the ◄►arrow key to More and press Enter. With Version
blinking, press Enter. The screen displays the manufacturer, model, and
software version information.
4.3.7 Showing Negative Oxygen Readings
For software version 1.4.4 or later, the instrument only displays
oxygen readings that are positive or zero. The instrument can be
reconfigured to show negative readings if sensor output drifts below
zero. This situation may arise after the instrument has been zeroed, as
time progresses the sensor may drift below the zero calibration setpoint.
To show negative oxygen readings on the display:
Press the System key.
TRAK/HLD Auto-Cal
PSWD Logout More
Use the Right or Left arrow keys and select More. Press Enter.
Version Self-Test
Show_Negative=NO
Use the Right or Left Arrow keys and select “Show_Negative=NO”.
Use the Up or Down key to toggle from NO to YES.
Press the Escape key twice to return to the analyze mode.
Teledyne Analytical Instruments
42
Trace Oxygen Analyzer
Operation
This preference is stored in non-volatile memory, so this
configuration is remembered after a power shutdown. If the instrument
is cold started, it will go back to default (not showing negative oxygen
readings).
4.4 The Zero and Span Functions
Note: Zeroing is not required in order to achieve the published
accuracy specification of this unit.
Zeroing will eliminate offset error contributed by sensor,
electronics, and internal and external sampling system and
improve performance beyond published specification limits.
The analyzer is calibrated using zero and span gases.
Any suitable oxygen-free gas can be used for zero gas as long as it
is known to be oxygen free and does not react adversely with the sample
system.
Although the instrument can be spanned using air, a span gas with a
known oxygen concentration in the range of 70–90% of full scale of the
range of interest is recommended. Since the oxygen concentration in air
is 20.9% (209,000 ppm), the cell can take a long time to recover if the
instrument is used for trace oxygen analysis immediately following
calibration in air.
Connect the calibration gases according to the instructions given in
Section 3.4.1, Gas Connections, observing all the prescribed
precautions.
Shut off the gas pressure before connecting it to the analyzer,
and be sure to limit the pressure to 40 psig or less when turning it
back on.
Readjust the gas pressure into the analyzer until the flowrate (as
read on the analyzer’s SLPM flowmeter) settles between 0.15 and 2.4
SLPM (approximately 0.2 - 5 SCFH).
If you are using password protection, you will need to enter your
password to gain access to either of these functions. Follow the
instructions in Sections 4.3.3 to enter your password. Once you have
gained clearance to proceed, you can enter the Zero or Span function.
Teledyne Analytical Instruments
43
Operation
3000TA- EU
4.4.1 Zero Cal
The Zero button on the front panel is used to enter the zero
calibration function. Zero calibration can be performed in either the
automatic or manual mode. In the automatic mode, an internal
algorithm compares consecutive readings from the sensor to determine
when the output is within the acceptable range for zero. In the manual
mode, the operator determines when the reading is within the acceptable
range for zero. Make sure the zero gas is connected to the instrument. If
you get a CELL FAILURE message skip to Section 4.4.1.3.
4.4.1.1 AUTO MODE ZEROING
Press Zero to enter the zero function mode. The screen allows you
to select whether the zero calibration is to be performed automatically or
manually. Use the ▲/▼ arrow keys to toggle between AUTO and MAN
zero settling. Stop when AUTO appears, blinking, on the display.
Zero: Settling: AUTO
To Begin
Press Enter to begin zeroing.
####PPM Zero
Slope=####
ppm/s
The beginning zero level is shown in the upper left corner of the
display. As the zero reading settles, the screen displays and updates
information on Slope (unless the Slope starts within the acceptable zero
range and does not need to settle further).
Then, and whenever Slope is less than 0.08 for at least 3 minutes,
instead of Slope you will see a countdown: 5 Left, 4 Left, and so forth.
These are five steps in the zeroing process that the system must
complete, AFTER settling, before it can go back to Analyze.
####PPM Zero
4 Left=###
ppm/s
Teledyne Analytical Instruments
44
Trace Oxygen Analyzer
Operation
The zeroing process will automatically conclude when the output is
within the acceptable range for a good zero. Then the analyzer
automatically returns to the Analyze mode.
4.4.1.2 MANUAL MODE ZEROING
Press Zero to enter the Zero function. The screen that appears
allows you to select between automatic or manual zero calibration. Use
the ▲/▼ keys to toggle between AUTO and MAN zero settling. Stop
when MAN appears, blinking, on the display.
Zero: Settling: Man
To Begin
Press Enter to begin the zero calibration. After a few seconds the
first of five zeroing screens appears. The number in the upper left hand
corner is the first-stage zero offset. The microprocessor samples the
output at a predetermined rate. It calculates the differences between
successive samplings and displays the rate of change as Slope= a value
in parts per million per second (ppm/s).
####ppm Zero
Slope=####
ppm/s
Note: It takes several seconds for the true Slope value to display.
Wait about 10 seconds. Then, wait until Slope is
sufficiently close to zero before pressing Enter to finish
zeroing.
Generally, you have a good zero when Slope is less than 0.05
ppm/s for about 30 seconds. When Slope is close enough to zero, press
Enter. In a few seconds, the screen will update.
Once span settling completes, the information is stored in the
microprocessor, and the instrument automatically returns to the Analyze
mode.
4.4.1.3 CELL FAILURE
Cell failure in the 3000TA-EU is usually associated with inability
to zero the instrument down to a satisfactorily low ppm reading. When
this occurs, the 3000TA-EU system alarm trips, and the LCD displays a
failure message.
Teledyne Analytical Instruments
45
Operation
3000TA- EU
#.# ppm Anlz
CELL FAIL/ ZERO HIGH
Before replacing the cell:
• Check your span gas to make sure it is within
specifications.
• Check for leaks downstream from the cell, where oxygen
may be leaking into the system.
If there are no leaks and the span gas is OK, replace the cell as described in Chapter 5, Maintenance.
4.4.2 Span Cal
The Span button on the front panel is used to span calibrate the
analyzer. Span calibration can be performed using the automatic mode,
where an internal algorithm compares consecutive readings from the
sensor to determine when the output matches the span gas concentration.
Span calibration can also be performed in manual mode, where the
operator determines when the span concentration reading is acceptable
and manually exits the function.
4.4.2.1 AUTO MODE SPANNING
Press Span to enter the span function. The screen that appears
allows you to select whether the span calibration is to be performed
automatically or manually. Use the ▲/▼ arrow keys to toggle between
AUTO and MAN span settling. Stop when AUTO appears, blinking, on
the display.
Span: Settling: AUTO
For Next
Press Enter to move to the next screen.
Calib. Holding time
Cal hold: 5 min
Teledyne Analytical Instruments
46
Trace Oxygen Analyzer
Operation
This menu allows the operator to set the time the analyzer should
be held in the span mode, after the readings of the analyzer settle. Five
minutes is the default, but it could be adjusted anywhere from 1 to 60
minutes by using the UP or DOWN keys.
Press Enter to move to the next screen.
Span Val: 000008.00
Span Mod #
Use the ▲/▼ arrow keys to enter the oxygen-concentration mode.
Use the ◄►arrow keys to blink the digit you are going to modify. Use
the ▲/▼ arrow keys again to change the value of the selected digit.
When you have finished typing in the concentration of the span gas you
are using (209000.00 if you are using air), press Enter to begin the Span
calibration.
#### ppm
Span
Slope=####
ppm/s
The beginning span value is shown in the upper left corner of the
display. As the span reading settles, the screen displays and updates
information on Slope. Spanning automatically ends when the span
output corresponds, within tolerance, to the value of the span gas
concentration. Then the instrument automatically returns to the analyze
mode.
4.4.2.2 MANUAL MODE SPANNING
Press Span to start the Span function. The screen that appears
allows you to select whether the span calibration is to be performed
automatically or manually.
Span: Settling:MAN
For Next
Teledyne Analytical Instruments
47
Operation
3000TA- EU
Use the ▲/▼ keys to toggle between AUTO and MAN span
settling. Stop when MAN appears, blinking, on the display. Press Enter
to move to the next screen.
Press Enter to move to the next screen.
Calib. Holding time
Cal hold: 5 min
This menu allows the operator to set the time the analyzer should
be held in the auto span mode. It does not affect anything in Manual
Mode. Just press Enter to continue.
Span Val: 000008.00
Span Mod #
Press ▲ () to permit modification (Mod #) of span value.
Use the arrow keys to enter the oxygen concentration of the span
gas you are using (209000.00 if you are using air). The ◄►arrows
choose the digit, and the ▲/▼ arrows choose the value of the digit.
Press Enter to enter the span value into the system and begin the
span calibration.
Once the span has begun, the microprocessor samples the output at
a predetermined rate. It calculates the difference between successive
samplings and displays this difference as Slope on the screen. It takes
several seconds for the first Slope value to display. Slope indicates rate
of change of the Span reading. It is a sensitive indicator of stability.
#### % Span
Slope=####
ppm/s
When the Span value displayed on the screen is sufficiently stable,
press Enter. (Generally, when the Span reading changes by 1% or less
of the full scale of the range being calibrated for a period of ten minutes
it is sufficiently stable.) Once Enter is pressed, the Span reading
changes to the correct value. The instrument then automatically enters
the Analyze function.
Teledyne Analytical Instruments
48
Trace Oxygen Analyzer
Operation
4.4.3 Span Failure
The analyzer checks the output of the cell at the end of the span. If
the raw output of the cell is less than 0.5 uA/ppm O2, the span will not
be accepted. The analyzer will return to the previous calibration values,
trigger the System Alarm, and display in the VFD:
Span Failed!!
This message will be shown for five seconds and the instrument
will return to the Analyze mode. In the upper right hand corner of the
VFD display “FCAL” will be shown. This message flag will help the
operator troubleshoot in case calibration was initiated remotely. To reset
the alarm and the flag message, the unit must be turned off by cycling
the standby key . It will reset if the next span cycle is correct.
A trace cell is unlikely to fail span. As explained before, when the
sensor reaches the end of its useful life, the zero offset begins to rise
until the analyzer finds the zero unsatisfactory. Nevertheless, feeding the
wrong span gas or electronics failure could set this feature off at the end
of the span. Consider this before replacing the cell.
4.5 The Alarms Function
The Model 3000TA-EU is equipped with 2 fully adjustable concentration alarms and a system failure alarm. Each alarm has a relay with a
set of form “C" contacts rated for 3 amperes resistive load at 250 V ac.
See Figure in Chapter 3, Installation and/or the Interconnection Diagram
included at the back of this manual for relay connections.
The system failure alarm has a fixed configuration described in
Chapter 3 Installation.
The concentration alarms can be configured from the front panel as
either high or low alarms by the operator. The alarm modes can be set as
latching or non-latching, and either failsafe or non-failsafe, or, they can
be defeated altogether. The setpoints for the alarms are also established
using this function.
Decide how your alarms should be configured. The choice will
depend upon your process. Consider the following four points:
1. Which if any of the alarms are to be high alarms and which if
any are to be low alarms?
Setting an alarm as HIGH triggers the alarm when the
oxygen concentration rises above the setpoint. Setting an
Teledyne Analytical Instruments
49
Operation
3000TA- EU
alarm as LOW triggers the alarm when the oxygen
concentration falls below the setpoint.
Decide whether you want the alarms to be set as:
• Both high (high and high-high) alarms, or
• One high and one low alarm, or
• Both low (low and low-low) alarms.
2. Are either or both of the alarms to be configured as failsafe?
In failsafe mode, the alarm relay de-energizes in an alarm
condition. For non-failsafe operation, the relay is energized
in an alarm condition. You can set either or both of the
concentration alarms to operate in failsafe or non-failsafe
mode.
3. Are either of the alarms to be latching?
In latching mode, once the alarm or alarms trigger, they will
remain in the alarm mode even if process conditions revert
back to non-alarm conditions. This mode requires an alarm to
be recognized before it can be reset. In the non-latching
mode, the alarm status will terminate when process
conditions revert to non- alarm conditions.
4. Are either of the alarms to be defeated?
The defeat alarm mode is incorporated into the alarm circuit
so that maintenance can be performed under conditions
which would normally activate the alarms.
The defeat function can also be used to reset a latched alarm.
(See procedures, below.)
If you are using password protection, you will need to enter your
password to access the alarm functions. Follow the instructions in
section 4.3.3 to enter your password. Once you have clearance to
proceed, enter the Alarm function.
Press the Alarm button on the front panel to enter the Alarm
function. Make sure that AL–1 is blinking.
AL—1
AL—2
Choose Alarm
Teledyne Analytical Instruments
50
Trace Oxygen Analyzer
Operation
Set up alarm 1 by moving the blinking over to AL–1 using the
◄►arrow keys. Then press Enter to move to the next screen.
AL—1 1000
ppm HI
Dft—N Fs—N Ltch—N
Five parameters can be changed on this screen:
• Value of the alarm setpoint, AL–1 #### ppm (oxygen)
• Out-of-range direction, HI or LO
• Defeated? Dft–Y/N (Yes/No)
• Failsafe? Fs–Y/N (Yes/No)
• Latching? Ltch–Y/N (Yes/No).
To define the setpoint, use the ◄►arrow keys to move the
blinking over to AL–1 ####. Then use the ▲/▼ arrow keys to change the
number. Holding down the key speeds up the incrementing or
decrementing. (Remember, the setpoint units are ppm O2.)
To set the other parameters use the ◄►arrow keys to move the
blinking over to the desired parameter. Then use the ▲/▼ arrow keys to
change the parameter.
Once the parameters for alarm 1 have been set, press Alarms again,
and repeat this procedure for alarm 2 (AL–2).
To reset a latched alarm, go to Dft– and then press either two
times or two times. (Toggle it to Y and then back to N.)
–OR –
Go to Ltch– and then press either ▲ two times or ▼ two times.
(Toggle it to N and back to Y.)
A
V
4.6 The Range Function
The Range function allows the operator to program up to three
concentration ranges to correlate with the DC analog outputs. If no
ranges are defined by the user, the instrument defaults to:
Low = 0–100 ppm
Med = 0–1,000 ppm
High = 0–1 0,000 ppm
Teledyne Analytical Instruments
51
Operation
3000TA- EU
The Model 3000TA-EU is set at the factory to default to
autoranging. In this mode, the microprocessor automatically responds to
concen-tration changes by switching ranges for optimum readout
sensitivity. If the current range limits are exceeded, the instrument will
automatically shift to the next higher range. If the concentration falls to
below 85% of full scale of the next lower range, the instrument will
switch to that range. A corresponding shift in the DC percent-of-range
output, and in the range ID outputs, will be noticed.
The autoranging feature can be overridden so that analog output
stays on a fixed range regardless of the oxygen concentration detected. If
the concentration exceeds the upper limit of the range, the DC output
will saturate at 1 VDC (20 mA at the current output).
However, the digital readout and the RS-232 output of the
concentration are unaffected by the fixed range. They continue to read
accurately with full precision. See Front Panel description in Chapter 1.
The automatic air calibration range is always 0-25 % and is not programmable.
4.6.1 Setting the Analog Output Ranges
To set the ranges, enter the range function mode by pressing the
Range button on the front panel.
L—100
M—1000
H—1 0000 Mode—AUTO
Use the ◄►arrow keys to blink the range to be set: low (L),
medium (M), or high (H).
Use the ▲/▼ arrow keys to enter the upper value of the range (all
ranges begin at 0 ppm). Repeat for each range you want to set. Press
Enter to accept the values and return to Analyze mode. (See note
below.)
Note: The ranges must be increasing from low to high, for
example, if range 1 is set as 0–100 ppm and range 2 is set
as 0–1,000 ppm, range 3 cannot be set as 0– 500 ppm
since it is lower than range 2.
Ranges, alarms, and spans are always set in ppm units (over the
entire 0-250,000 ppm range), even though all concentration-data outputs
Teledyne Analytical Instruments
52
Trace Oxygen Analyzer
Operation
change from ppm units to percent when the concentration is above
10,000 ppm.
4.6.2 Fixed Range Analysis
The autoranging mode of the instrument can be overridden, forcing
the analyzer DC outputs to stay in a single predetermined range.
To switch from autoranging to fixed range analysis, enter the range
function by pressing the Range button on the front panel.
Use the ◄►arrow keys to move the blinking over AUTO.
Use the ▲/▼ arrow keys to switch from AUTO to FX/LO, FX/M ED,
or FX/H I to set the instrument on the desired fixed range (low, medium,
or high).
L—100
M—1000
H—1 0000 Mode—FX/ LO
or
L—100
M—1000
H—1 0000 Mode—FX/MED
or
L—100
M—1000
H—1 0000 Mode—FX/ HI
Press Escape to re-enter the Analyze mode using the fixed range.
Note: When performing analysis on a fixed range, if the oxygen
concentration rises above the upper limit (or default value)
as established by the operator for that particular range, the
output saturates at 1 VDC (or 20 mA). However, the digital
readout and the RS-232 output continue to read the true
value of the oxygen concentration regardless of the analog
output range.
Teledyne Analytical Instruments
53
Operation
3000TA- EU
4.7 The Analyze Function
Normally, all of the functions automatically switch back to the
Analyze function when they have completed their assigned operations.
Pressing the Escape button in many cases also switches the analyzer
back to the Analyze function. Alternatively, you can press the Analyze
button at any time to return to analyzing your sample.
4.8 Signal Output
The standard Model 3000TA-EU Trace Oxygen Analyzer is
equipped with two 0–1 VDC analog output terminals accessible on the
back panel (one concentration and one range ID), and two isolated 4–20
mA DC current outputs (one concentration and one range ID).
See Rear Panel in Chapter 3, Installation, for illustration.
The signal output for concentration is linear over the currently
selected analysis range. For example, if the analyzer is set on range that
was defined as 0–100 ppm O2, then the output would be:
ppm O2
0
10
20
30
40
50
60
70
80
90
100
Voltage Signal
Output (VDC)
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
Current Signal
Output (mA dc)
4.0
5.6
7.2
8.8
10.4
12.0
13.6
15.2
16.8
18.4
20.0
The analog output signal has a voltage which depends on the
oxygen concentration AND the currently activated analysis range. To
relate the signal output to the actual concentration, it is necessary to
know what range the instrument is currently on, especially when the
analyzer is in the autoranging mode.
To provide an indication of the range, a second pair of analog
output terminals are used. They generate a steady preset voltage (or
Teledyne Analytical Instruments
54
Trace Oxygen Analyzer
Operation
current when using the current outputs) to represent a particular range.
The following table gives the range ID output for each analysis range:
Range
Voltage (V)
Current (mA)
LO
0.25
8
MED
0.50
12
HI
0.75
16
CAL (0-25%)
1.00
20
IMPORTANT: In the event of loss of flow through the analyzer, if the vent
is vented to a location of high oxygen content, oxygen will
back diffuse through the vent line and in most cases
quickly saturate the cell with oxygen which can then
require a quite long purge down time for the sensor when
then exposed to low oxygen concentrations. In the event
that flow is to be interrupted into the analyzer, it is
suggested that the user do one of the following:
1.
Bag the sensor in nitrogen during this time
2.
Install a shut off valve on the vent port of the
analyzer or somewhere within the users sample
system.
4.9 Maintenance Schedule
The Maintenance function offers the user the ability to set a screen
notification after a user-defined period that the sensor needs
replacement. For instance, after installing a new sensor, the user can set
a defined interval (6 months is the default setting) after which a
notification on the Analyze screen will advise that sensor maintenance is
needed. This function can be set so that in addition to the notification, it
also triggers the system failure alarm. It can also be turned off.
To access the Maintenance function use the Use the ◄►arrow
keys to navigate to the fourth System menu.
Maintenance Serial#
Stream SerMode More
Teledyne Analytical Instruments
55
Operation
3000TA- EU
Use the RIGHT button to move the blinking to Maintenance and
then press ENTER to go to the next screen.
Set: M: 0 D: 7 On
Nxt: M: 0 D: 6 Reset
The timer in the maintenance function is set in months (M) and
days (D) on the top line of the display. The maximum setting is M=30
and D=30. The default is M=6 and D=0. The user has the option of
turning the function ON or OFF or to have it trigger the system alarm
(Alarm).
To change the notification period, use the RIGHT arrow key to
move the blinking to the appropriate field M(onth) or D(ay) and then use
the UP/DOWN keys to increment or decrement the value. Press ENTER
when the appropriate setting is displayed.
The second line is the actual countdown and displays the time left
before a notification or alarm will occur. It counts down until the
Maintenance warning is set. To re-start the counter the operator must
select “Reset” and press ENTER. The operator is prompted to press
ENTER again to re-affirm the restart of the counter.
Reset Maint Timeout?
Yes No
Pressing ESC will return you to the previous screen.
There are three modes to the Maintenance counter: Off, On, and
Alarm.
OFF:
ON:
Alarm:
Turns off the counter, thus Maintenance mode is off.
Turns on the counter but at the end of the countdown “MAINT”
is shown on the display but the System alarm contacts are not
triggered.
Turns ON the counter and will display “MAINT” on the
Analyze screen at the end of the countdown plus trigger the
system alarm contacts.
Teledyne Analytical Instruments
56
Trace Oxygen Analyzer
Operation
In this mode, when the counter reaches M:0 and D:0, the system
alarm contact on the back of the instrument is triggered and
the upper right corner of the display in Analyze mode will
have the normal “Anlz”replaced by “MAINT”.
Note: The system alarm can be reset only after the timer has
been reset.
0.00
ppm MAINT
Range: 0 - 100
4.10 Sensor Detection
The sensor detection feature detects whether a sensor is installed in
the analyzer or not. The software checks if the output of the amplifier is
within a user-defined threshold around the electronic zero (output of the
amplifier without sensor). In the event that the sensor reads a near zero
(or below the user-defined threshold) ADC count for 30 minutes it will
trigger a system failure alarm with a corresponding message on the
Analyze screen.
In this function, the user can set the count threshold below which
the alarm will occur as well as reset the 30 minute timer back to 30
minutes.
Note: The 30 minute delay is set at the factory and cannot be
modified.
The default threshold is set to 300 but can be increased or
decreased by the user using this function. The sensor fail function can
also be turned OFF or ON within this function.
To access the Sensor function, use the LEFT/RIGHT arrow keys to
navigate to the last System menu.
Sensor
Press ENTER. The next screen will appear:
Teledyne Analytical Instruments
57
Operation
3000TA- EU
Sensor FAIL: ON
Thresh: 200 Reset
In this screen the featured can be turned on and off on the first line.
The Sensor Detection feature is ON by default. To turn the alarm ON or
OFF, use the LEFT/RIGHT arrow keys to move the blinking to ON (or
OFF). Press the UP or DOWN keys to toggle between ON and OFF.
Press ENTER to save your choice.
The detection window (Thresh) around the electronic zero level is
set on the second line. It defaults to 300 but it can be adjusted. This
numbers is the ADC counts on the A to D converter and it is equivalent
to 0.114% of the full scale of the A to D converter.
Adjust the threshold for the alarm by moving the blinking over to
the value adjacent to Thres:. Use the UP/DOWN keys to increase or
decrease the threshold value. When the desired value is displayed, press
ENTER to save it.
The fixed 30 minute timer begins to count down to zero whenever
the ADC sensor count falls below the set threshold. If within the
countdown, the user sets a new and higher threshold or replaces the
sensor to remedy the near zero output, the countdown timer can be reset
back to 30 minutes manually using the Reset feature.
To reset the counter to 30, use the arrow keys to move the blinking
to Reset and then press the ENTER key.
The timer and the ADC counts can be displayed on the Analyze
screen by pressing the ESC key twice.
6.09
ppm
Anlz
1- 136730 30:00
The second line will display three numbers for a few seconds. The
first number (1) is the gain of the amplifier (1 is the highest gain used
for low ppm analysis).
The second number (136730) is the ADC count from the A to D
converter.
The third number (30:00) is the timer in minutes: seconds, and this
is the value that it shows when it is not triggered.
Teledyne Analytical Instruments
58
Trace Oxygen Analyzer
Operation
When the timer reaches 00:00, the system alarm is triggered and the
message “FSEN“ and “ERRS” appear in the upper right corner of the
display instead of “Anlz”.
0.01
ppm
FSEN
Range: 0 - 100
0.01
ppm
ERRS
Range: 0 - 100
To reset the message and the system alarm contacts, the operator
must go back to the Sensor function in the system menu and select
Reset, then press the ENTER key. The display will be followed by a
prompt to confirm the reset.
Reset Sensor Fail?
Yes No
4.11 Valve Box Functions
Three additional functions are available from the System Menu
that are used exclusively with analyzers that interface with a Teledyne
VB Valve Box ot Teledyne Adapter Profibus (TAPA) board. These are:
• Serial #
• Stream
• SerMode (Serial Mode)
If your analyzer is not connected to a VB style valve box or have a
TAPA interface installed, these functions will have no effect.
4.11.1 Serial#
At this screen, you can set the serial number of the analyzer so that
it is passed on to a valve box through the RS232. The default value is
100000. The serial number can be found on a sticker either just inside
opening the door of the analyzer or on the rear panel.
Teledyne Analytical Instruments
59
Operation
3000TA- EU
To enter the serial number navigate to the third System menu using
the LEFT/RIGHT keys and selecting More until the following display
appears:
Maintenance Serial#
Stream SerMode More
Use the LEFT/RIGHT keys again to move to Serial# and press
ENTER. The Enter Serial Number screen will appear.
Enter Serial#
100000
Use the UP/DOWN keys to enter the serial number and then press
ENTER to save.
4.11.2 Stream
The Stream function is useful when the analyzer is connected with
a TAI Valve Box. It allows the user to switch the input stream from the
analyzer rather than the Valve Box. This assumes that the analyzer is in
local mode and not being controlled by an external source. If the
analyzer is not interfaced with a Valve Box, this function has no input.
To enter the Stream function navigate to the third System menu
using the LEFT/RIGHT keys and selecting More until the following
display appears:
Maintenance Serial#
Stream SerMode More
Use the LEFT/RIGHT keys again to move to Stream and press
ENTER. The display will change to the following:
Teledyne Analytical Instruments
60
Trace Oxygen Analyzer
Operation
Select Sample Stream
1
Press
Use the UP/DOWN keys to display the proper valve to select (1,2
or 3). Then Press ENTER. This will cause the valve controlling that
stream to open.
4.11.3 SerMode
There are two serial communication modes available in the Model
3000TA-EU: Profibus and standard RS232 (default).
The Profibus mode is used primarily in conjunction with a TAI
Valve Box. It is used to communicate specific commands for controlling
or receiving input from the Valve Box. This feature requires either
special software or a Profibus Adapter Board (TAPA) available from
TAI which converts the RS232 to Profibus and adds additional
commands.
To view or change the current communication mode, use the
LEFT/RIGHT and ENTER keys to navigate to the third System menu
and select SerMode from the menu.
Serial Mode: Profi
or
Serial Mode: Std (RS232)
Use the UP/DOWN keys to toggle between Profi (Profibus) and Std
(RS232), and then press ENTER.
The RS232 defaults to support connectivity to Teledyne’s Valve
Box or Teledyne Adapter Profibus (TAPA) board.
Note that valve box or TAPA board support requires the baud rate
to be set to 9600. Dip switches in motherboard must be set as shown
below to change baud rate to 9600:
Position 1 set to ON
Position 2 to OFF
Position 3 to 8 set to OFF.
Teledyne Analytical Instruments
61
Operation
3000TA- EU
Teledyne Analytical Instruments
62
Trace Oxygen Analyzer
Teledyne Analytical Instruments
Operation
63
Maintenance
3000TA- EU
Maintenance
5.1 Routine Maintenance
Aside from normal cleaning and checking for leaks at the gas
connections, routine maintenance is limited to replacing Micro-Fuel
cells and fuses, and recalibration. For recalibration, see Section 4.4
Calibration.
Warning: See warnings on the title page of this manual.
5.2 Cell Replacement
The L-2 Micro-Fuel Cell is a sealed electrochemical transducer
with no electrolyte to change or electrodes to clean. When the cell
reaches the end of its useful life, it is replaced. The spent fuel cell should
be discarded according to local regulations. This section describes fuel
cell care as well as when and how to replace it.
5.2.1 Storing and Handling Replacement Cells
To have a replacement cell available when it is needed, it is recommended that one spare cell be purchased 9-10 months after
commissioning the 3000TA-EU, or shortly before the end of the cell's
one year warranty period.
CAUTION:
DO NOT STOCKPILE CELLS. THE WARRANTY
PERIOD STARTS ON THE DAY OF SHIPMENT.
The spare cell should be carefully stored in an area that is not
subject to large variations in ambient temperature (75°F nominal, 24°C)
or to rough handling.
WARNING:
The sensor used in the model 3000TA-EU Trace
Oxygen Analyzer uses electrolytes which contain
toxic substances, mainly Lead and potassium
Teledyne Analytical Instruments
64
Trace Oxygen Analyzer
Maintenance
hydroxide, that can be harmful if touched,
swallowed, or Inhaled. Avoid contact with any fluid
or powder in or around the unit. What may appear to
be plain water could contain one of these toxic
substances. In case of eye contact, immediately
flush eyes with water for at least 15 minutes. Call
physician. (See appendix, Material Safety Data
Sheet.)
CAUTION:
DO NOT DISTURB THE INTEGRITY OF THE CELL
PACKAGE UNTIL THE CELL IS TO ACTUALLY BE
USED. IF THE CELL PACKAGE IS PUNCTURED AND
AIR IS PERMITTED TO ENTER, THE CELL WILL
REQUIRE AN EXCESSIVELY LONG TIME TO REACH
ZERO AFTER INSTALLATION (1-2 WEEKS!).
5.2.2 When to Replace a Cell
The characteristics of the Micro-Fuel Cell show an almost constant
output throughout its useful life and then fall off sharply towards zero at
the end. Cell failure in the 3000TA-EU is usually characterized inability
to zero the instrument down to a satisfactorily low ppm reading. When
this occurs, the 3000TA-EU system alarm trips, and the LCD displays a
failure message.
#.# ppm Anlz
CELL FAIL/ ZERO HIGH
Before replacing the cell:
a) Check your span gas to make sure it is within
specifications.
b) Check for leaks downstream from the cell, where oxygen
may be leaking into the system.
If there are no leaks and the span gas is OK, replace the cell.
5.2.3 Removing the Micro-Fuel Cell
The Micro-Fuel cell is located inside the stainless steel cell block
behind the front panel (see Figure 5-1). To remove an existing cell:
1. Remove power to the instrument by unplugging the power
cord at the power source.
Teledyne Analytical Instruments
65
Maintenance
3000TA- EU
2. Open the front panel door by pressing the release button on
the top right corner of the door all the way in with a narrow
gauge tool, such as a small screwdriver, and releasing it.
3. With one hand placed underneath the cell block ready to
catch the Micro-Fuel cell, lift up on the stainless steel gate in
front of the cell block. This releases the cell and cell holder
from the block. The cell and holder will fall out in your hand.
WARNING: Risk of electric shock high voltage exposed at
the end of enclosure!
Figure 5-1: Removing the Micro-Fuel
Teledyne Analytical Instruments
66
Trace Oxygen Analyzer
Maintenance
5.2.4 Installing a New Micro-Fuel Cell
It is important to minimize the amount of time that a Teledyne
Trace Oxygen Sensor is exposed to air during the installation process.
The quicker the sensor can be installed into the unit, the faster your TAI
O2 sensor will recover to low O2 measurement levels. The procedure for
installing a cell depends on the type of cell your instrument is using.
Section 5.2.4.1 describes the procedure to use for all cells except the
Insta-Trace cell. The Insta-Trace cell installation procedure is given in
Section 5.2.4.2.
5.2.4.1 STANDARD TRACE OXYGEN SENSOR CELL INSTALLATION
This section describes the procedures for removing and installing a
conventional trace oxygen sensor such as the A2C, L2C, or B2C.
CAUTION:
DO NOT TOUCH THE SENSING SURFACE OF THE
CELL. IT IS COVERED WITH A DELICATE TEFLON
MEMBRANE THAT CAN LEAK WHEN PUNCTURED.
THE SENSOR MUST BE REPLACED IF THE
MEMBRANE IS DAMAGED.
Before installing a new cell, check the O-ring in the base of the cell
holder. Replace if worn or damaged.
Place the cell on the holder with the screen side facing down.
Note: There is a small location hole drilled in the holder. This
hole mates with a guide pin on the bottom rear of the cell
block. The hole in the cell block holder must align with the
guide pin on the cell block.
1. Remove power from instrument.
2. Remove the old sensor (if installed) from the analyzer.
3. Purge the analyzer at approximately 1 SCFH flow rate with
N2 (or zero gas with the sensor holder removed).
4. Remove sensor from double bag storage.
5. Remove sensor shorting button.
6. Place sensor on sensor holder so that the gold contact plate of
the sensor is facing up towards the sky.
7. Install sensor and sensor holder into cell block. If the A2C or
B2C sensor is used, a cell adapter must be used between the
sensor and the sensor holder (P/N B66378).
Teledyne Analytical Instruments
67
Maintenance
3000TA- EU
8. With O-ring in place, align the guide pin with the hole on the
cell holder. Then, with the holder, lift cell into the cell block.
9. Push the gate on the cell block down so that the slots on the
side of the gate engage the locating screws on the side of the
block. This forces the holder into position and forms a gastight seal.
10. Purge system with sample or zero gas.
11. Power-up.
Note: If steps 4 through 10 are accomplished quickly (elapsed
time less than 15 seconds), recovery to less than 1ppm
level should occur in less than 8 hours.
5.2.4.2 INSTA-TRACE CELL INSTALLATION
The installation procedure for the Insta-Trace family of cells is
slightly different. The major difference is that this cell arrives with a thin
plastic film over the sensing surface that gets punctured during the
installation process. This film is used for maintaining a reasonable cell
shelf life as well as protecting the cell during shipping.
To install an Insta-Trace Cell:
1. Remove power from instrument.
2. Initiate a nitrogen or zero gas purge flow at approximately 1
SCFH. Maintain the flow through the entire installation
process. This flow is required to minimize the cell’s exposure
to air during installation. Without a purge gas flowing, the
lifetime of the cell would be severely compromised plus the
analyzer would require a long wait period until an
appropriate zero level of the cell could be attained.
3. Remove the old sensor (if installed) from the analyzer by
lifting up the lever on the cell block with one hand under the
block to catch the old cell as it falls from the holder.
4. With the sensor holder removed, initiate a nitrogen or zero
gas purge flow at approximately 1 SCFH. Note that the
sensor holder has a locating hole which is designed to fit onto
a pin inside the cell block. This aligns the sensor contacts
with spring loaded contacts inside the cell block.
5. Remove sensor from double bag storage.
6. Remove sensor shorting button which covers the contacts on
the top of the sensor.
Teledyne Analytical Instruments
68
Trace Oxygen Analyzer
Maintenance
7. With the gold contact plate of the sensor facing up, push the
sensor up into the cell block.
8. While holding the sensor in position in the block place the
cell holder into the block and align the hole on the holder
with the pin on the block.
9. When the holder and pin are aligned, a slight push on the
holder will puncture the thin protective film on the cell.
10. Push the gate on the cell block down so that the slots on the
side of the gate engage the locating screws on the side of the
block. This forces the holder into position and forms a gastight seal.
11. Continue purging the block and cell with nitrogen or zero gas
for an additional 1 to 2 hours.
12. Power-up, the analyzer is now ready for calibration and
service.
5.3 Fuse Replacement
1. Disconnect the AC power and remove the AC plug from the
rear panel of the instrument.
2. Place small screwdriver in the notch of the fuse block, and
pry the cover off as shown in Figure 5-2.
Figure 5-2: Removing Fuse Block from Housing
3. Within the fuse block cavity is a removable fuse clip which,
due to a locating pin in the cavity, can be placed in one of
Teledyne Analytical Instruments
69
Maintenance
3000TA- EU
two orientations, UP or DOWN. The UP position has the
jumper bar facing upwards. This corresponds to the
American fuse orientation. To change from American to
European fuses, remove the fuse clip and flip fuse clip over
180 degrees so that the jumper bar is on the bottom (DOWN)
similar to graphic in Figure 5-3.
Note: Figure 5-3 shows an older style fuse block, the newer
version is similar.
4.
Insert the proper fuse(s) into the clip.
Note: The European position requires two fuses whereas in the
American position, only one fuse is required.
5. Replace the clip making sure the notch on the fuse clip
engages the locating pin in the fuse block cavity.
6. Reassemble the housing as shown in Figure 5-2.
American Fuses
European Fuses
Figure 5-3: Installing Fuses
5.4 System Self Diagnostic Test
1. Press the System button to enter the system mode.
2. Use the ◄►arrow keys to move to More, and press Enter.
3. Use the ◄►arrow keys to move to Self-Test, and press
Enter. The following failure codes apply:
Table 5-1: Self Test Failure Codes
Power
0
1
2
3
OK
5 V Failure
15 V Failure
Both Failed
Teledyne Analytical Instruments
70
Trace Oxygen Analyzer
Maintenance
Analog
0
OK
1
DAC A (0–1 V Concentration)
2
DAC B (0–1 V Range ID)
3
Both Failed
Preamp
0
OK
1
Zero too high
2
Amplifier output doesn't match test input
3
Both Failed
5.5 Major Internal Components
The Micro-Fuel cell is accessed by unlatching and swinging open
the front panel, as described earlier. Other internal components are
accessed by removing the rear panel and sliding out the entire chassis.
See Figure 5-4, below. The gas piping is illustrated in Figure 2-4, and
the major electronic components locations are shown in Figure 2-5, in
chapter 2.
WARNING: See warnings on the title page of this manual.
The 3000TA-EU contains the following major components:
• Analysis Section
• Micro Fuel Cell (L-2 or other)
• Stainless steel cell block
• Sample system
• Power Supply
• Microprocessor
• Displays
• 5 digit LED meter
• 2 line, 20 character, alphanumeric, VFD display
Teledyne Analytical Instruments
71
Maintenance
3000TA- EU
•
RS-232 Communications Port.
See the drawings in the Drawings section in back of this manual for
details.
Figure 5-4: Rear-Panel Screws
To detach the rear panel, remove only the 14 screws marked with an X.
5.6 Cleaning
If instrument is unmounted at time of cleaning, disconnect the
instrument from the power source. Close and latch the front-panel access
door. Clean outside surfaces with a soft cloth dampened slightly with
plain clean water. DO NOT use any harsh solvents such as paint thinner
or benzene.
For panel-mounted instruments, clean the front panel as prescribed
in the above paragraph. DO NOT wipe front panel while the instrument
is controlling your process.
5.7 Troubleshooting
Problem: Erratic readings of the Oxygen concentration as reported by
the analyzer.
Possible Cause:
Teledyne Analytical Instruments
72
Trace Oxygen Analyzer
Maintenance
The analyzer may have been calibrated in an inaccurate fashion.
Solution:
Turn the analyzer off, then back on again. Press the System key when
prompted by the analyzer “Press System for default Values”. This will
return the analyzer to its default settings in calibration and zero values.
If erratic behavior continues replace the sensor.
Possible Cause:
Atmospheric Oxygen may be diffusing in through the vent and affecting
the oxygen level which the sensor sees.
Solution:
Increase flow rate and/or length or vent tubing in order to dilute or minimize the diffusion of oxygen from the vent back to the sensor.
Problem: Inaccurate zero operation (i.e. the user has zeroed the
analyzer accidentally on gas much higher than one would normally use
for a zero gas).
Solution:
Turn the analyzer off, then back on again. Press the System key when
prompted by the analyzer "Press System for default Values". This will
return the analyzer to its default settings in calibration and zero values.
Now proceed to carefully calibrate and zero the analyzer.
Teledyne Analytical Instruments
73
Appendix
3000TA- EU
Appendix
A-1 Model 3000TA-EU Specifications
Packaging: General Purpose
•
Flush panel mount (Standard).
•
Sample System:
Ranges:
Alarms:
Displays:
Digital Interface:
Power:
Operating Temperature:
Humidity:
Accuracy:
Relay rack mount. Contains either
one or two instruments in one 19"
relay rack mountable plate
(Optional).
Sensor: Teledyne L-2 trace analysis
Micro-Fuel Cell. Cell Block: 316 stainless
steel.
All wetted parts of 316 stailess steel. 90 %
Response Time: 65 seconds at 25 °C (77
°F).
Three user definable ranges from 0–10
ppm to 0–250,000 ppm, plus air
calibration range 0- 250,000 ppm (25 %).
Autoranging with range ID output.
One system-failure alarm contact to detect
power failure or sensor-zero failure.
Two adjustable concentration threshold
alarm contacts with fully programmable
setpoints.
Two-line by 20-character, VFD screen,
and one 5 digit LED display.
Full duplex RS-232 communications port.
Universal power supply 85-250 VAC, at
47-63 Hz, 0.9 A MAX.
0-50 °C (32-122 °F)
99% Altitude: 1,609 m
±2% of full scale at constant temperature.
±5% of full scale over operating
Teledyne Analytical Instruments
74
Trace Oxygen Analyzer
Appendix
temperature range (except 0-10 ppm
analysis range) once thermal equilibrium
is reached.
±1 ppm on 0-10 ppm analysis range, once
thermal equilibrium is reached.
Analog Outputs: 0-1 VDC percent-of-range,
0-1 VDC range ID
4-20 mA DC—isolated—percent-ofrange,
4-20 mA DC—isolated—range ID.
Dimensions: 19 cm high, 24.9 cm wide, 31 cm deep
(6.96 in high, 8.7 in wide, 12.2 in deep).
A-2 Recommended 2-Year Spare Parts List
Qty
Part Number
Description
1
1
1
1
2
C65507A
C62371
C62368-A
C73870-A
F 1296
1
1*
1
O165
Back Panel Board
Front Panel Board
Trace Preamplifier Board
Main Computer Board
Fuse, 2A, 250 V 5x20 mm Slow
Blow
O-ring
Micro-Fuel Cell
Restrictor Kit
A68729
* See page iii for sensor used in this instrument
Note: Orders for replacement parts should include the part
number (if available) and the model and serial number of
the instrument for which the parts are intended.
Teledyne Analytical Instruments
75
Appendix
3000TA- EU
Orders should be sent to:
TELEDYNE Analytical Instruments
16830 Chestnut Street
City of Industry, CA 91748
Phone (626) 934-1500, Fax (626) 961-2538
Web: www.teledyne-ai.com
or your local representative.
Teledyne Analytical Instruments
76
Trace Oxygen Analyzer
Appendix
A-3 Drawing List
D66316
D77800
Final Assembly/Outline Drawing
Final Assembly Trace Oxygen Analyzer 3000TA-EU
Series
A-4 19-inch Relay Rack Panel Mount
Figure A-1: Single and Dual 19" Rack Mounts
(dimensions in mm)
Teledyne Analytical Instruments
77
Appendix
3000TA- EU
A.5 Application notes
3000 SERIES ANALYZERS
APPLICATION NOTES ON PRESSURES AND FLOW
RECOMMENDATIONS
The 3000 series analyzers require reasonably regulated sample
pressures. While the 3000 analyzers are not sensitive to variations of
incoming pressure provided they are properly vented to atmospheric
pressure. The pressure must be maintained as to provide a useable flow
rate trough the analyzer. Any line attached to sample vent should be 1/4
or larger in diameter.
FLOW RATE RECOMMENDATIONS:
A usable flow rate for a 3000 series analyzer is one which can be
measured on the flowmeter. This is basically .2-2.4 SLPM. The
optimum flow rate is 1 SLPM (mid scale).
Note: Response time is dependent on flow rate, a low flow rate
will result in slow response to O2 changes in the sample
stream. The span flow rate should be the approximately
same as the sample flow rate.
CELL PRESSURE CONCERNS:
The sensors used in 3000 series analyzers are optimized to function
at atmospheric pressure. At pressures other than atmospheric the
diffusion rate of O2 will be different than optimum value. Higher
pressures will produce faster O2 diffusion rates resulting in higher O2
reading and shorter cell life. To use a 3000 series analyzer at a cell
pressure other than atmospheric, the analyzer must be calibrated with a
known calibration gas at the new cell pressure to adjust for the different
diffusion rate. Cell pressures below 2/3 atmospheric are not
recommended because they tend to cause excessive internal expansion
which may result in seal failure.
For operation at cell pressures other than atmospheric, care must be
taken not to change the sample pressure rapidly or cell damage may
occur. For cell pressures above atmospheric, caution must be exercised
to avoid over pressuring the cell holder. (Percent analyzers will require
some type of cell retainer to prevent the cell from being pushed out by
Teledyne Analytical Instruments
78
Trace Oxygen Analyzer
Appendix
the pressure). For operation at pressures below atmospheric pressure a
suffix C (clamped) cell is required.
RESTRICTION DEVICES:
For proper operation, all 3000 series analyzers require a flow
restriction device. This device is typically a restrictor or a valve. This
restriction device serves two functions in the sample path. The first
function is to limit the flow rate of the sample through the analyzer. A
restrictor is chosen to operate over a range of pressures and provide a
useable flow rate over that range.
The second function that the restriction device provides is a
pressure drop. This device is selected to provide the only significant
pressure drop in the sample path.
RESTRICTOR KIT
The current revision of the 3000 series analyzers are supplied with
a kit containing two restrictors and a union which are user installed.
These parts supplied to give the end user more flexibility when installing
the analyzer. The restrictor kit is suitable for high and low positive
pressure applications as well as vacuum service (atmospheric pressure
sample) applications (see manual for installation instructions).
The standard restrictor (BLUE DOT) is recommended for pressures
between 5 psig and 50 psig. For positive low pressure application (5 psig
or less) the un-marked restrictor is better suited . For non-pressurized
sample applications the marked restrictor should be used and configured
for vacuum service.
For extremely low positive pressure applications (less than 2 psig)
the vacuum service configuration should provide higher performance
(higher flow rates). For vacuum service the end user must supply a
vacuum pump and a bypass valve for the pump. A vacuum level of 5 10 inches of mercury should provide the optimum flow rate.
CAUTION:
FLOW RESTRICTORS HAVE VERY SMALL ORIFICES
AND MAY BE PLUGGED BY SMALL PARTICLES
(.005” DIA OR LARGER) A SAMPLE FILTER MUST
BE INCLUDED IN THE SAMPLE LINE PRIOR TO THE
RESTRICTOR! A 60 MICRON FILTER IS
RECOMMENDED.
3000TA-EU EXAMPLES:
Teledyne Analytical Instruments
79
Appendix
3000TA- EU
Example 1: With an incoming pressure of 10 psig the standard restrictor
(blue dot) will provide a flow rate of .76 SLPM. Up-stream of the
restrictor the sample line pressure will be 10 psig, while down stream
(including the cell) the pressure will be at atmospheric pressure.
(analyzer vented to atmospheric pressure) Note, all other pressure drops
in the sample path are insignificant at these flowrates. This insures that
the cell operates at atmospheric pressure. At very high flow rates (off
scale of flow-meter), pressure drops other than the restriction device
could become significant, and result in pressurizing the cell.
Example 2: A 3000TA-EU is configured for vacuum service as follows.
The un-marked restrictor is placed in the sample vent port. The
downstream end of the restrictor is then connected to a vacuum pump
and bypass valve. The bypass valve is adjusted to provide a flow rate of
1 SLPM. The sample pressure between the pump and the restrictor will
be approximately -7 inches of mercury, while the pressure in the balance
of the sample system including the cell will be approximately at
atmospheric pressure. (Provided the sample flow into the analyzer is not
blocked.)
BYPASS:
To improve the system response, a bypass can be added to increase
the sample flow rate to the analyzer by a factor of ten. A bypass
provides a sample flow path around the analyzer of 2-18 scfh, typically.
CALIBRATION GAS:
For 3000 series analyzers r with the Auto-Cal option, the customer
must supply control valves (or restrictors) for any SPAN or ZERO gas
source which is attached to the Auto-Cal ports. The valve should be
adjusted to the same flow rate as the sample gas. When restrictors are
used, the gas pressure must be adjusted to achieve the proper flow rate.
OPERATION WITHOUT A RESTRICTORDEVICE:
Operation without a restrictor device is not recommend as
mentioned above. A 3000TA-EU without any flow restrictor device was
tested on 11-19-97. This results in a flow rate of 2.4 slpm at 1 psig. This
is a cv of 0.023 for the standard sample system.
Teledyne Analytical Instruments
80
Trace Oxygen Analyzer
Appendix
REFERENCE: FLOW_1 .XLS & FLOW _2.XLS for
information on flow rates at various pressures.
TAI PART NUMBERS
Restrictor Kit:
Union (ss)
LP. Restrictor
Std. Restrictor
Nut
Ferrule
Ferrule
A68729
U11
R2323
R2324
N73
F73
F74
(low pressure /vac. service )
Blue Dot
Both ferrules are required
CONVERSIONS:
1 PSI
1 SCFH
=
=
2.04 INCHES OF MERCURY (in. Hg.)
0.476 SLPM
Teledyne Analytical Instruments
81
Appendix
3000TA- EU
A-5 Material Safety Data Sheet
Section I - Product Identification
Product Name: Micro-fuel Cells
Mini-Micro-fuel Cells
Super Cell, all classes except T-5F
Electrochemical Oxygen Sensors, all classes
Manufacturer: Teledyne Electronic Technologies
Analytical Instruments
Address: 16380 Chestnut Street,
City of Industry, CA 91749
Phone: (626) 961-9221
Technical Support: (626) 934-1673
Environment, Health and (626) 934-1592
Safety:
Date Prepared: 11/23/98
Section II - Physical and Chemical Data
Chemical and Common Potassium Hydroxide (KOH), 15% (w/v)
Names: Lead (Pb), pure
CAS Number: KOH 1310-58-3
Pb 7439-92-1
KOH (15% w/v)
Pb (pure)
Melting Point/Range: -10 to 0 °C
328 °C
Boiling Point/Range: 100 to 115 °C
1744 °C
Specific Gravity: 1.09 @ 20 °C
pH: >14
Solubility in Water: Completely soluble
Percent Volatiles by Vol.: None
Appearance and Odor: Colorless, odorless
solution
11.34
N/A
Insoluble
N/A
Grey metal,
odorless
Teledyne Analytical Instruments
82
Trace Oxygen Analyzer
Appendix
Section III -Physical Hazards
Potential for fire and explosion: The electrolyte in the Micro-fuel Cells is not
flammable. There are no fire or explosion hazards associated with Micro-fuel
Cells.
Potential for reactivity: The sensors are stable under normal conditions of
use. Avoid contact between the sensor electrolyte and strong acids.
Section IV - Health Hazard Data
Primary route of entry: Ingestion, eye/skin contact
Exposure limits: OSHA PEL: 0.05 mg./cu.m. (Pb)
ACGIH TLV: 2 mg/ cu.m. (KOH)
Effects of overexposure
Ingestion: The electrolyte could be harmful or fatal if
swallowed.
Oral LD50 (RAT) = 3650 mg/kg
Eye: The electrolyte is corrosive; eye contact could
result in permanent loss of vision.
Dermal: The electrolyte is corrosive; skin contact could
result in a chemical burn.
Inhalation: Liquid inhalation is unlikely.
Signs/symptoms of exposure: Contact with skin or eyes will cause a burning
sensation and/or feel soapy or slippery to
touch.
Medical conditions
aggravated by exposure: None
Carcinogenicity: NTP Annual Report on Carcinogens: Not
listed
LARC Monographs: Not listed
OSHA: Not listed
Other health hazards: Lead is listed as a chemical known to the State
of California to cause birth defects or other
reproductive harm.
Teledyne Analytical Instruments
83
Appendix
3000TA- EU
Section V - Emergency and First Aid Procedures
Eye Contact: Flush eyes with water for at least 15 minutes
and get immediate medical attention.
Skin Contact: Wash affected area with plenty of water and
remove contaminated clothing. If burning
persists, seek medical attention.
Ingestion: Give plenty of cold water. Do not induce
vomiting. Seek medical attention. Do not
administer liquids to an unconscious person.
Inhalation: Liquid inhalation is unlikely.
Section VI - Handling Information
NOTE:
The oxygen sensors are sealed, and under normal circumstances, the
contents of the sensors do not present a health hazard. The following
information is given as a guide in the event that a cell leaks.
Protective clothing: Rubber gloves, chemical splash goggles.
Clean-up procedures: Wipe down the area several times with a wet
paper towel. Use a fresh towel each time.
Protective measures Before opening the bag containing the sensor
during cell replacement: cell, check the sensor cell for leakage. If the
sensor cell leaks, do not open the bag. If there
is liquid around the cell while in the
instrument, put on gloves and eye protection
before removing the cell.
Disposal: Should be in accordance with all applicable
state, local and federal regulations.
NOTE:
The above information is derived from the MSDS provided by the
manufacturer. The information is believed to be correct but does not
purport to be all inclusive and shall be used only as a guide.
Teledyne Analytical Instruments shall not be held liable for any
damage resulting from handling or from contact with the above
product.
Teledyne Analytical Instruments
84