Endress + Hauser Operating Instructions Liquiline M CM42
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Operating Instructions
Liquiline M CM42
Two-wire transmitter for pH/ORP measurement with
analog sensors
Part 1: Commissioning
BA381C/07/en/POD
Valid as of:
Software version 10.04.00
Overview of documentation supplied
Operating Instructions
The Operating Instructions are split into two parts:
• Commissioning
– Description of your device
– All steps that you have to carry out once only during initial commissioning
• Operation
– Software description
– Troubleshooting
– All steps that you have to carry out during the routine operation of your device,
depending on your measuring task
!
Note!
This part of the Operating Instructions is the documentation on commissioning.
CD-ROM
Contents:
• Fundamentals of pH and conductivity measurement
• Communication
• Overview of analytical measuring technology
• Accessories
• Safety
• Ordering information
In addition
• 1 manufacturer's certificate
• 1 test certificate to EN 10204-3.1
• Adhesive labels:
– Replacement nameplate (landscape format)
– Wiring adhesive label
Liquiline M CM42
Table of contents
1
Safety instructions . . . . . . . . . . 4
9
Accessories . . . . . . . . . . . . . . . 40
1.1
1.2
1.3
1.4
1.5
1.6
Designated use . . . . . . . . . . . . . . . . . . . . . .
Incoming acceptance, transport, storage . . .
Installation, commissioning and operation .
Operational safety . . . . . . . . . . . . . . . . . . .
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Notes on safety conventions and icons . . . .
9.1
9.2
9.3
Installation kits . . . . . . . . . . . . . . . . . . . . . 40
Weather protection . . . . . . . . . . . . . . . . . . 40
Communication . . . . . . . . . . . . . . . . . . . . 40
2
Identification . . . . . . . . . . . . . . 6
2.1
2.2
2.3
2.4
Device designation . . . . . . . . . . . . . . . . . . .
Scope of delivery . . . . . . . . . . . . . . . . . . . .
Reordering documentation . . . . . . . . . . . . .
Certificates and approvals . . . . . . . . . . . . . .
3
Device description . . . . . . . . . 10
3.1
3.2
Housing closed . . . . . . . . . . . . . . . . . . . . . 10
Housing open . . . . . . . . . . . . . . . . . . . . . . 10
4
Installation . . . . . . . . . . . . . . . 11
4.1
4.2
4.3
4.4
System setup . . . . . . . . . . . . . . . . . . . . . .
Installation conditions . . . . . . . . . . . . . . .
Installation instructions . . . . . . . . . . . . . .
Post-installation check . . . . . . . . . . . . . . .
5
Wiring . . . . . . . . . . . . . . . . . . 17
5.1
5.2
Electrical connection . . . . . . . . . . . . . . . . 17
Post-connection check . . . . . . . . . . . . . . . 27
6
Commissioning. . . . . . . . . . . . 28
6.1
6.2
6.3
6.4
6.5
6.6
Function check . . . . . . . . . . . . . . . . . . . .
Configuring the device address . . . . . . . . .
Device master and type files . . . . . . . . . . .
Quick Setup (local operation) . . . . . . . . . .
User administration . . . . . . . . . . . . . . . . .
Calibration . . . . . . . . . . . . . . . . . . . . . . . .
7
Maintenance. . . . . . . . . . . . . . 35
8
Troubleshooting - hardware . . 36
8.1
8.2
8.3
8.4
Replacing modules . . . . . . . . . . . . . . . . . .
Spare parts . . . . . . . . . . . . . . . . . . . . . . . .
Return . . . . . . . . . . . . . . . . . . . . . . . . . . .
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . .
4
4
4
5
5
5
6
8
8
9
10 Technical data . . . . . . . . . . . . . 41
10.1
10.2
10.3
10.4
10.5
10.6
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Output . . . . . . . . . . . . . . . . . . . . . . . . . . .
Electrical connection . . . . . . . . . . . . . . . . .
Performance characteristics . . . . . . . . . . . .
Environment . . . . . . . . . . . . . . . . . . . . . .
Mechanical construction . . . . . . . . . . . . . .
41
41
42
42
43
44
Index. . . . . . . . . . . . . . . . . . . . 45
Endress+Hauser
11
12
14
16
28
28
29
29
32
33
36
37
39
39
Safety instructions
Liquiline M CM42
1
Safety instructions
1.1
Designated use
Liquiline CM42 is a two-wire transmitter for liquid analysis in all areas of process engineering:
• Chemical processes
• Pharmaceutical industry
• Food technology
If the device is used for any purpose other than that described, this poses a threat to the safety
of people and the entire measuring system and is thus not permitted.
The manufacturer does not accept liability for damage caused by improper or non-designated
use.
1.2
Incoming acceptance, transport, storage
• Make sure the packaging is not damaged!
Inform your supplier of any damage to the packaging.
Please keep the damaged packaging until any issues have been resolved.
• Make sure the contents are not damaged!
Inform your supplier of any damage to the contents.
Please keep the damaged goods until any issues have been resolved.
• Check the scope of delivery against the delivery papers and your order to ensure it is complete
and nothing is missing.
• Pack the product in such a way as to protect it reliably against impact and moisture for storage
and transportation. Optimum protection is provided by the original packaging materials. In
addition, the permitted ambient conditions must be observed (see Technical data).
• If you have any queries, please contact your supplier or local sales center.
1.3
Installation, commissioning and operation
Observe the following points:
• Installation, commissioning, operation and maintenance of the measuring system must only
be carried out by trained technical personnel.
The technical personnel must be authorized to perform the tasks by the owner-operator.
• The electrical connection may only be established by an electrical technician.
• The technical personnel must have read and understood these Operating Instructions and
must follow the instructions they contain.
• Prior to commissioning the entire measuring point, check that all connections are correct.
Make sure that electric cables and hose connections are not damaged.
• Do not commission damaged products. Protect them against unintentional startup. Label and
identify the damaged product as defective.
• Faults at the measuring point may only be rectified by authorized and properly trained
personnel.
4
Endress+Hauser
Liquiline M CM42
Safety instructions
• If the faults cannot be eliminated, take the products out of service and protect them against
unintentional startup.
• Repairs not described in these Operating Instructions may only be carried out directly at the
manufacturer's or by the Service Organization.
1.4
Operational safety
The transmitter is designed to meet state-of-the-art safety requirements, has been tested and left
the factory in a condition in which it is safe to operate.
The applicable regulations and European standards have been taken into account.
As the user, you are responsible for observing the following safety regulations:
• Installation guidelines
• Local standards and regulations
1.5
Return
If returning the transmitter, please send it cleaned to your sales center.
Use the original packaging when returning the device.
1.6
Notes on safety conventions and icons
1.6.1 Warnings
#
Warning!
This symbol alerts you to dangers.
If the warning is not observed, this can result in serious injury to people or damage
to property.
"
Caution!
This symbol alerts you to possible malfunctions resulting from incorrect operation.
If not observed, this could result in damage to property.
!
Note!
This symbol indicates important items of information.
1.6.2 Document symbols
È ä 1 This symbol stands for a cross-reference to a certain page (e.g. Page 1).
È å 2 This symbol stands for a cross-reference to a certain graphic (e.g. Fig. 2).
Endress+Hauser
5
Identification
Liquiline M CM42
2
Identification
2.1
Device designation
2.1.1 Nameplate
Nameplates can be found:
• On the outside of the housing (lasered)
• On the packaging (adhesive label, portrait format)
• On the adhesive label sheet provided (landscape format, for your own use)
Compare the data on the nameplate with your order.
The nameplate provides you with the following
information on your device:
1. Order code
2. Serial number
3. Device version
4. Measuring range(s)
5. Input, output values
6. Approvals (depending on the version ordered)
7. Protection class
8. Ambient temperature range
9. Approval symbols
a0003281
Fig. 1: Nameplate information
10. Safety notices, warnings
2.1.2 Serial number and order code
The order code and serial number of your device can be found in the following locations:
• On the nameplate
• On the front page of the Operating Instructions
• In the delivery papers
!
6
Note!
• The documentation only contains information that was valid at the time of delivery.
• If you have any queries, please contact your local sales center.
Endress+Hauser
Liquiline M CM42
Identification
2.1.3 Product structure
CM42-
!
Sensor input
A
No module
C
Conductivity, conductive measurement, analog sensor
I
Conductivity, inductive measurement, analog sensor
K
Digital sensor: conductivity, conductive measurement
M
Digital sensor: pH/ORP measurement, glass sensor
N
Digital sensor: pH measurement, ISFET sensor
O
Digital sensor: amperometric oxygen measurement
P
pH (glass/ISFET) or ORP, analog sensor
Approval
A
Non-hazardous area, CSA GP
G
Atex II (1)2 G EEx ia/ib IIC T6 / II 3 D
H
Atex II 3G EEx nA II T6
N
NEPSI Ex nA II T6
P
FM IS NI Cl. I, Div. 1&2, Groups A-D
S
CSA IS NI Cl. I, II, III, Div. 1&2, Groups A-G
Certificate
A
Not selected
B
Test certificate to EN 10204, 3.1
C
Test certificate to EN 10204, 3.1, factory calibration certificate
Output
0
1 current output 4 to 20 mA, HART®
1
2 current outputs 4 to 20 mA, 1x HART®
2
PROFIBUS PA
3
FOUNDATION Fieldbus
Housing
0
Plastic housing
1
Stainless steel housing 1.4404 (AISI 316L)
Cable entry
0
M20 x 1.5
1
NPT ½"
2
G½
Software
EA Standard version
EB Advanced version
EC Advanced Plus version
Device language
D
De/En (Docs, German)
E
En/De (Docs, English)
F
Fr/En (Docs, French)
S
Es/En (Docs, Spanish)
Z
Zh/En (Docs, Chinese)
Documentation
0
Installation + operation
Additional fittings
0
Basic version
Complete order code
Note!
Please ask your sales office for other device versions, e.g. other software or documentation
languages.
Endress+Hauser
7
Identification
2.2
•
•
•
•
•
•
•
Liquiline M CM42
Scope of delivery
1 transmitter in the version ordered
1 wall securing unit incl. 4 flat head screws
1 sheet of adhesive labels (nameplates, terminal connection diagrams)
1 test certificate to EN 10204-3.1
1 Operating Instructions Part 1, BA381C "Commissioning"
1 Operating Instructions Part 2, BA382C "Operation"
1 CD-ROM with additional documentation
If you have any queries, please contact your supplier or local sales center.
2.3
Reordering documentation
What to order?
• Reordering lost documentation or documentation of which several copies are required:
You must quote the serial number of your device.
You receive the same documentation as received with the delivery.
• First time ordering appropriate documentation for devices which you have assembled yourself
by installing a module:
You have to indicate the order code the device corresponds to after installing the module.
Specify a serial number also.
You receive the documentation to suit your newly modified device.
How to order?
• By Internet (www.endress.com/liquiline-documentation → å 2):
You receive PDF files of both parts of the Operating Instructions.
• Through your sales center:
Indicate you want the documentation as hard copy.
a0006062
Fig. 2: www.endress.com/liquiline-documentation
8
Endress+Hauser
Liquiline M CM42
2.4
Identification
Certificates and approvals
With this declaration, the manufacturer guarantees that the product conforms to the regulations
of European Directive 89/336/EEC. This is proven by observing the standards listed in the
declaration of conformity.
Endress+Hauser
9
Device description
Liquiline M CM42
3
Device description
3.1
Housing closed
1
1
2 2
3
4
5-8
Wall securing plate
Alarm LED
Display
Navigator
Soft keys,
assignment depends on menu
3
4
8
7
6
5
a0001067
Fig. 3: CM42 with closed display cover
3.2
Housing open
a0002448
Fig. 4:
1
2
3
10
With open display cover (without wiring)
DAT memory module
CPU module
Sensor module
4
5
6
DIN rail
Threaded joints
CDI socket (service interface)
Endress+Hauser
Liquiline M CM42
Installation
4
Installation
4.1
System setup
A complete system consists of:
• Liquiline M CM42
• An immersion assembly, e.g. CPA140 or flow assembly, e.g. CPA240 or retractable assembly,
e.g. CPA475; each with or without a potential matching pin (PML)
• A pH/ORP sensor: e.g. CPS71 (pH glass) or CPS471 (ISFET)
• A measuring cable: CPK9 (pH) or CPK12 (ISFET)
• A programmable logic controller (PLC) with Profibus interface incl. PC and operating
software, segment coupler and Profibus-PA terminating resistor
a0001642
Fig. 5: Example of a measuring system
1
PC with operating software
2
PLC
3
Segment coupler
4
Liquiline M CM42
Endress+Hauser
5
6
7
8
Special measuring cable
Retractable assembly CPA475 (manual version)
pH/ORP sensor
Terminating resistor
11
Installation
4.2
Liquiline M CM42
Installation conditions
4.2.1 Dimensions
a0001074
Fig. 6: Dimensions
12
Endress+Hauser
Liquiline M CM42
Installation
4.2.2 Mounting plate
a0001075
Fig. 7: Mounting plate
4.2.3 Weather protection cover
Note!
When installing outside, always use the weather protection cover (see 'Accessories', CD-ROM)
to protect the transmitter against rain and snow.
100°
230 (9.01)
107 (4.21)
!
90°
253 (9.96)
98 (3.86)
224 (8.82)
98 (3.86)
151 (5.94)
mm (inch)
a0001671
Fig. 8: Weather protection cover
Endress+Hauser
13
Installation
Liquiline M CM42
4.3
Installation instructions
4.3.1 Wall or field mounting
There are many ways of mounting the device:
• Wall mounting
• Mounting on a vertical pipe or post (round or square)
• Mounting on railing or on a horizontal pipe (round or square)
!
Note!
You require the post mount to mount on a pipe or post. This is an accessory and is not included
in the scope of delivery (see "Accessories").
1
2
1
2
4
3
3
4
a0002162
14
a0003092
Fig. 9: Mounting on a vertical pipe/post
Fig. 10: Side view
1
2
3
4
CM42 (as delivered, incl. a mounting plate)
Securing screws (part of the post mount, see Accessories)
Mounting plate (post mount)
Pipe or post (round/square)
Endress+Hauser
Liquiline M CM42
Installation
1
2
1
2
4
3
3
4
a0002164
a0003093
Fig. 11: Mounting on a horizontal pipe/railing
Fig. 12: Side view
1
2
3
4
CM42 (as delivered)
Securing screws (post mount)
Mounting plate (post mount)
Pipe or railing
a0002166
Fig. 13: Wall mounting (optional weather protection cover)
4.3.2 Panel mounting
!
Note!
• For panel mounting, you need the installation kit consisting of tensioning screws and a front
seal. This is an accessory (see "Accessories") and is not included in the scope of delivery.
• If installing the units above one another, you must observe a minimum distance for the
cable glands of the upper device.
• If installing the units beside one another, you must observe a minimum distance for opening
the front of the housing.
• If arranging in a square, you must take into account the lengths of the mounting plates at the
rear of the device or the cable glands for minimum spacing distance.
Endress+Hauser
15
Installation
Liquiline M CM42
a0002402
Fig. 14: Panel mounting: left side view, right front view
4.4
Post-installation check
• After installation, check the transmitter for damage.
• Check whether the transmitter is protected against moisture and direct sunlight.
16
Endress+Hauser
Liquiline M CM42
#
Wiring
5
Wiring
5.1
Electrical connection
Warning!
• The electrical connection may only be established by an electrical technician.
• The electrical technician must have read and understood these Operating Instructions and
must follow the instructions they contain.
• Prior to beginning any wiring work, make sure voltage is not applied to any of the cables.
5.1.1 Opening the housing
"
1.
Loosen the four screws on the front with a Phillips screwdriver:
2.
Open the housing:
Caution!
Do not use any sharp or pointed objects, such as a screwdriver, knife etc., to open the housing.
Otherwise, you might damage the housing seal, scratch the housing etc.
Endress+Hauser
17
Wiring
Liquiline M CM42
5.1.2 Housing grounding
"
Caution!
You must connect the "fixing plate" to the foundation ground with a separate ≥2.5 mm2 (i14
AWG) functional ground.
a0003617
Fig. 15: Grounding the housing
1
2
Fixing plate
≥2.5 mm2 (i14 AWG) functional ground
5.1.3 Cable grounding in housing
!
Note!
• If possible, only use terminated genuine cables.
• Ground the sensor cable in the housing of the transmitter as shown in the following diagrams
(sample cable, does not necessarily correspond to the original cable).
1
2
3
a0002679
18
a0002680
Fig. 16: Terminated cable (example)
Fig. 17: Inserting the cable
1
2
3
4
Outer shield (exposed)
Cable cores with ferrules
Cable sheath (insulation)
Grounding clip
a0002681
Fig. 18: Grounding the cable
(functional ground)
Endress+Hauser
Liquiline M CM42
Wiring
5.1.4 Cable terminals
1. Insert a suitable screwdriver into the opening
of the terminal spring (square opening) until
the stop.
a0003602
Fig. 19: Opening the terminal
2. Insert the terminated cable end into the
terminal opening (round opening).
a0003603
Fig. 20: Inserting the cable
3. Remove the screwdriver. Make sure the cable
is securely positioned in the terminal.
a0003604
Fig. 21: Releasing the terminal spring
5.1.5 Signal output / power supply
Permitted supply voltage
!
Profibus PA
9 to 32 V DC
Bus current consumption:
22 mA
Note!
Use a fieldbus cable grounded on both sides (device and PCS).
There are three possible ways of connecting:
1.
Two-wire cabling shielded on both sides, "hard grounding"
(generally to be given priority over "capacitive connection to ground")
2.
Shielded two-wire cabling, "capacitive connection to ground" (shield grounded on device
side via capacitor, "C-module" accessory required)
Use if there is the risk of large potential equalization currents.
3.
Use the fieldbus connection socket
Endress+Hauser
19
Wiring
Liquiline M CM42
Version 1: "Hard grounding"
• Position the cable shielding on the "fixing plate".
• Connect the cable cores as per the assignment (→ å 22).
a0004061
Fig. 22: View in device (CPU module)
a0001640
Fig. 23: Wiring diagram
Version 2:: "Capacitive connection to ground"
• Pull back the braided shield, put the stranded extension wire of the C-module (item 1) onto
the exposed shield and tighten the clip:
• Position the extension wire on the "fixing plate".
• Connect the cable cores as per the assignment (→ å 24).
20
Endress+Hauser
Liquiline M CM42
Wiring
a0004072
Fig. 24: View in device (CPU module)
a0004074
Fig. 25: Wiring diagram
Version 3: "Fieldbus connection socket" (accessory):
• Screw the fieldbus connection socket into the housing bushing.
• Trim the connection cores of the socket to approx. 15 cm.
• Connect the cable cores as per the assignment. In doing so, you must position the cable
shielding (GN/YE) on the "fixing plate" (→ å 26).
a0002364
Fig. 26: View in device (CPU module)
Endress+Hauser
a0004063
Fig. 27: Wiring diagram
21
Wiring
Liquiline M CM42
5.1.6 Sensor connection
Explanation of abbreviations in the following diagrams:
!
22
Abbreviation
Meaning
pH
Ref
Src
Drn
PM
ϑ
d.n.c.
Signal from pH membrane glass
Signal from reference electrode
Source
Drain
Potential matching
Signal of the temperature sensor
Do not connect!
Note!
• Cable colors indicated as per IEC 757 (see CD–ROM).
• You must connect shielded connections and terminals with functional ground ()) (there is
no protective ground (*) for plastic housings).
Endress+Hauser
Liquiline M CM42
Wiring
With PM (symmetrical)
Glass electrodes
CM42
112
113
111
CPK9
Sensor
GN
J
WH
YE
315
CPK1
316
320
318
BN
BK
317
PM
Ref
pH
319
a0001088
Fig. 28: View in device (sensor module)
Endress+Hauser
a0001079
Fig. 29: Wiring diagram
23
Wiring
Liquiline M CM42
ISFET sensors
CM42
112
113
111
CPK12
Sensor
GN
WH
J
YE
315
Src
Ref
316
320
318
RD
BN
Drn
PM
BK
317
319
a0001090
Fig. 30: View in device (sensor module)
24
a0001076
Fig. 31: Wiring diagram
Endress+Hauser
Liquiline M CM42
Wiring
Pfaudler electrodes
Pfaudler electrode, absolute
Type 03/type 04/type Reiner
Pfaudler electrode, relative
Type 18/type 40
a0006051
Fig. 32: Wiring diagram
Endress+Hauser
a0006052
Fig. 33: Wiring diagram
25
Wiring
Liquiline M CM42
Without PM (asymmetrical)
Glass electrodes
CPK9 / CPK1
CM42
112
Sensor
GN
J
113
111
WH
YE
315
CPK1
316
320
BK
318
Ref
317
pH
319
d.n.c.
BN
a0001089
Fig. 34: View in device (sensor module)
26
a0001080
Fig. 35: Wiring diagram
Endress+Hauser
Liquiline M CM42
Wiring
ISFET sensors
CPK12
CM42
112
113
111
Sensor
GN
WH
J
YE
Src
315
Ref
316
320
RD
Drn
BK
318
d.n.c.
317
319
BN
a0001084
Fig. 36: View in device (sensor module)
5.2
a0001077
Fig. 37: Wiring diagram
Post-connection check
Device state and specifications
1.
Are the transmitter and cables free from damage on the outside?
Electrical connection
2.
Are the mounted cables strain relieved?
3.
Are the cables run without loops and cross-overs?
4.
Are the signal lines correctly connected in accordance with the wiring diagram?
5.
Have you positioned unused connection wires on the "fixing plate"?
Endress+Hauser
27
Commissioning
#
#
Liquiline M CM42
6.
Are all the connection wires securely positioned in the cable terminals?
7.
Are all the cable entries installed, tightened and sealed?
Warning!
Only put the transmitter into operation if you can answer yes to all these questions. Otherwise,
safe operation cannot be guaranteed. In such instances, the manufacturer does not accept any
liability for the device.
6
Commissioning
6.1
Function check
Warning!
• Check that all connections have been established correctly.
• Make sure that the supply voltage matches the voltage indicated on the nameplate!
6.2
Configuring the device address
The address must always be configured for a PROFIBUS-PA device. If the address is not
configured correctly, the measuring device is not recognized by the process control system.
All devices are delivered from the factory with the address 126. You can use this address to
check the function of the device and to connect to a PROFIBUS-PA network.
You then have to change this address to integrate additional devices.
The following methods are available for configuring the address:
!
1.
By means of PROFIBUS communication
2.
By means of hardware settings (DIL switch).
Note!
• Valid device addresses are in the range from 0 to 126.
• Each address may only be assigned once in a PROFIBUS-PA network.
6.2.1 Configuring the device address by means of PROFIBUS communication
You configure the address by means of the "Set Slave Adr" service.
!
Note!
You can only configure the address via the bus if DIL switch 8 is set to "On" (= factory setting).
6.2.2 Configuring the device address by means of hardware (DIL switch)
To configure the device address, you must open the housing and change the keys of the DIL
switch on the CPU module.
You set the address (0 to 126) using keys 1-7. To do so, you must set key 8 to the "Off" position
(= hardware setting).
28
Endress+Hauser
Liquiline M CM42
Commissioning
You can set any number between 0 and 126 by using binary code, e.g.:
• 0:
• 1 (20=1):
• 18 (21+24=2+16=18):
6.3
Device master and type files
The device master file (GSD) is needed to configure a PROFIBUS-DP network. The GSD (simple
text file) contains information on the data transmission rate the device supports or what digital
information the PLC receives from the device and in what format.
!
Note!
Every device is given an ID number by the PROFIBUS User Organization (PNO). The name of
the device master file (GSD) is derived from this.
For Endress+Hauser, the ID number always starts with "15XX".
Device designation
ID No.
GSD
Bitmaps
Liquiline M CM42 pH
1543 Hex
EH3x1543.gsd
Eh1543_d.bmp
Eh1543_n.bmp
Eh1543_s.bmp
6.4
Quick Setup (local operation)
1.
Connect the supply voltage.
2.
Wait for the initialization to be complete.
3.
Completely work through the "Quick Setup" submenu in "PARAM".
6.4.1 Menu structure
Function name of display (local operation)
Bus parameter name
PARAM
Quick Setup
Endress+Hauser
Language
DISPLAY_LANGUAGE
Date format
DISPLAY_DATE_FORMAT
29
Commissioning
Liquiline M CM42
Function name of display (local operation)
Bus parameter name
Set date
Time format
DISPLAY_TIME_FORMAT
Set time
Measured value
SWITCH_OPERATING_MODE_CHANNEL_1
Potential matching
GEN_INSTALLATION_PAL
Sensor type pH
Internal buffer
(pH and sensor type="Glass")
SENSOR_TYPE_CH1
Temperature sensor
SENSOR_TYPE_TEMP
PH_SENSOR_INTERNAL_BUFFER_CH1
Temperature unit
DISPLAY_UNIT_TEMPERATURE
Buffer manufacturer (only pH)
PH_CAL_BUFFER_MANUFACTURER
Temp. compensation (only pH)
Medium temperature
(Temp. compensation="Man. comp.")
PH_CAL_TEMP_COMP_TYPE
PH_CAL_TEMP_COMP_REF_VALUE
6.4.2 Configuration options
The abbreviation "AC" is used in the table header. AC stands for "Access Controls", i.e. read and
write authorization.
The table only indicates the rights of the "Maintenance" user group as the "Specialist" group has
administrator status and thus always has full rights (--> "Commissioning/User administration").
• R = Read only, i.e. the maintenance engineer can only read the data
• R/W = Read/Write, i.e. as a maintenance engineer, you can also change values.
!
Note!
The access controls do not take effect until password protection has been activated!
Function
Options
AC
Info
Language
Options
• English
• Language ordered
R/W
Note!
!
"Language ordered" refers to the language that
you selected for your device using the order code
("device language").
Factory setting
Language ordered
Date format
Options
• DD.MM.YYYY
• MM.DD.YYYY
If you select the other language in question, all
the other settings remain intact.
R/W
Factory setting
DD.MM.YYYY
Set date
30
Depends on the format
DD.MM.YYYY
Editing mode:
DD (day): 1 to 31
MM (month): 1 to 12
YYYY (year): 2005 to 2100
R/W
Endress+Hauser
Liquiline M CM42
Commissioning
Function
Options
AC
Info
Time format
Options
• hhmmss (24 h)
• hhmmss (am / pm)
R/W
24-hour display or 12-hour display
Editing mode:
hh (hour): 0 to 23 / 0 am to 12 pm
mm (minutes): 0 to 59
ss (seconds): 0 to 59
Factory setting
hhmmss (24 h)
Set time
Depends on the format
hh:mm:ss
R/W
Measured value
Options
• pH
• ORP mV
R
Factory setting
pH
Potential matching
Options
• With PM
• Without PM
R
Indicate whether you want to measure
symmetrically (= with PM) or asymmetrically (=
without PM).
Note!
!
More information on symmetrical and
Factory setting
With PM
asymmetrical measurement on the CD-ROM.
Sensor type pH
Options
• Glass
• ISFET
R
Factory setting
Glass
Internal buffer
–2.0 to 16.0 pH
R
Only available if you have selected "Sensor
type" = "Glass".
R/W
If "Pt100/Pt1000" is selected, the system automatically detects whether a Pt100 or Pt1000 is
connected.
Factory setting
7.0
Temperature sensor
Options
• None
• Pt100 / Pt1000
Factory setting
Pt100 / Pt1000
Temperature unit
Options
• °C
• °F
R/W
Factory setting
°C
Buffer manufacturer
Options
• E+H (NIST)
• Ingold/Mettler
• DIN 19267
• DIN 19266
• Merck/Riedel
R/W
only pH
Factory setting
E+H (NIST)
Endress+Hauser
31
Commissioning
Liquiline M CM42
Function
Options
AC
Info
Temp. compensation
Options
• Off
• Auto comp. (ATC)
• Man. comp.
R/W
Menu not available if you have selected
"Measured value" = "ORP mV".
R/W
Only for temp. compensation = "Man.
comp."
Factory setting
Auto comp. (ATC)
Medium temperature
–50 to + 250 °C
(–58 to + 482 °F)
Factory setting
25 °C (77 °F)
6.5
User administration
To protect the device against unwanted or unauthorized access, you can assign an access code
to the menu.
!
Note!
The factory setting for the access code is an empty field (= no entry). The prompt to enter a
password is deactivated. Setting is possible for everyone.
Assigning/changing access codes:
1.
Go to PARAM menu, "General settings"/"User administration".
2.
Select the type of protection:
– No password protection (= factory setting).
– Password protection per code entry.
3.
If you select "Enter code", select the user role and assign a password to each of the user
roles.
6.5.1 User roles and access authorization
If the password protection has been defined and activated, you have to log in with the
appropriate user role before you can make changes to the settings in the menu.
32
Endress+Hauser
Liquiline M CM42
Commissioning
Here, you will be prompted by the device to enter an access code as soon as you want to
configure a value:
User roles and access authorization (factory setting):
!
Maintenance
Specialist
• Read and, sometimes, write access in the PARAM and
DIAG menus
• Read and write access in the calibration menu
Admin status:
• Read and write access in all menus
• Change the access authorization for other user groups
Note!
If you are logged in as the "Specialist" (= Administrator), you can change the access
authorization of all other user groups:
1.
Select a software function for which you want to assign authorization to, or take
authorization from, another user group.
2.
Press and hold down the navigator button for at least 3 seconds.
The display shows all the user groups and the current access authorization to the function
selected.
3.
Use the navigator to change the authorization in question and confirm with "OK".
4.
Make a note of the changes in your operating logbook or similar.
6.6
Calibration
Calibration is necessary:
• At initial commissioning
• After replacing a sensor
• After periods of standstill
• At sensible, process-dependent intervals1)
1)
Depending on the process conditions, the intervals can range from several times daily to once quarterly.
Endress+Hauser
33
Commissioning
Liquiline M CM42
6.6.1 Types of calibration
Select the calibration type in the menu "PARAM"/"Sensor..."/"Cal. Settings"/"Type of calibration" or in the Quick Setup.
• Two-point calibration
– With automatic buffer recognition (only pH):
You select the buffer type (e.g. DIN 19266). The device automatically recognizes the buffer
during calibration.
– With fixed buffer (only pH):
You select two buffer solutions from the menu.
– Manual:2)
You enter the buffer values during calibration.
• Single-point calibration
– You enter an offset.
• Numeric input:
– You enter the slope, zeropoint and temperature (only pH).
– You enter the mV value (for ORP).
• Grab sample calibration
– You take a sample of the medium. The device determines a measured value while you
analyze the sample in the laboratory.
Then you enter the laboratory measured value. The device calibrates on the basis of the
laboratory value.
6.6.2 Calibrate
Proceed as follows to calibrate the sensor:
1.
Go through the menu "Param"/"Quick Setup" completely3).
2.
Remove the sensor from the process.
3.
Clean the sensor.
4.
Press the soft key for "CAL".
5.
Follow the instructions in the menu.
6.
Finish calibrating by switching back to the measuring mode.
7.
Install the sensor back into the process.
Your measuring point is now ready for operation.
!
2)
3)
34
Note!
• If calibration is aborted using ESC, or if the calibration is faulty, the system continues to use
the original calibration data. A calibration error is shown as plain text on the display.
• Any offset set is automatically deleted after accepting the calibration.
Not available during initial commissioning if you are working with the factory settings.
Only for initial commissioning
Endress+Hauser
Liquiline M CM42
7
Maintenance
Maintenance
Clean the front of the housing with usual commercial cleaning agents.
The front is resistant to the following in accordance with DIN 42 115:
• Alcohol (briefly)
• Diluted acids (max. 2% HCl)
• Diluted alkalis (max. 3% NaOH)
• Soap-based household cleaner
"
Caution!
Never use any of the following for cleaning purposes:
• Concentrated mineral acids or alkalis
• Benzyl alcohol
• Methylene chloride
• High-pressure steam
Endress+Hauser
35
Troubleshooting - hardware
Liquiline M CM42
8
Troubleshooting - hardware
8.1
Replacing modules
Exchange modules when replacing defective modules or in the event of changes to the
configuration of your device.
1
1
2
a0004575
a0004577
Fig. 38: Removing module
Fig. 39: Housing guides
1
1
2
Removal aids
Guides on housing wall
Guides in DIN rail
1
a0004578
a0004579
Fig. 40: Module, left side (cable connections = front)
Fig. 41: Module, right side (cable connection = front)
1
1
Guides (e.g. CPU module to fit into the guides in
housing)
2
36
Guides (e.g. input module to fit into the right-hand
guides of the CPU module)
Removal aids
1.
Disconnect Liquiline from the power supply and open the housing.
2.
Remove the cable connections from the module you want to replace.
3.
Pull out the two removal aids on the module until the stop (→ å 38).
The module can now be easily removed from the DIN rail.
4.
Slide the new module into the guides (→ å 39, → å 40, → å 41).
5.
Push the two removal aids on the module in the direction of the DIN rail until the stop.
This locks the module onto the DIN rail.
6.
Connect the cables in accordance with the wiring diagram (see "Wiring").
7.
Close the housing, connect the sensor and check that the entire measuring system is
working correctly.
Endress+Hauser
Liquiline M CM42
8.2
Troubleshooting - hardware
Spare parts
a0004475
Fig. 42: Exploded view
Please refer to the following table for item names and order numbers for spare parts.
Endress+Hauser
37
Troubleshooting - hardware
Item
38
Liquiline M CM42
Kit CM42
Order number
1
1)
Housing base, plastic, G½
• Base
• Plate G½ (item 15)
• Fixing plate cpl. SS
• 3 threaded joints G½ (item 14)
71001363
2
CPU module, PA+FF, non-Ex
• FMPA1 insertion module (FC2W1+FBPA1), non-Ex
• Terminal strip (item 3)
51518003
3
CPU module terminal set, PA+FF
51517482
4
Ribbon cable for plastic housing
• 1 cable, conf. 40xAWG28 EX, 90 mm (3.5") crimp
51517503
5
Hinge for plastic housing
• Hinge cpl. CM42 KS
• 2 special screws K40x10 Torx self-form.
51517500
6
DAT chip with software
• FMDC1 FTC electronics BG EX
51517509
7
Top housing section, plastic, with display
• Display, keys, navigator, screws and cover plate
• Hinge (item 5)
• Ribbon cable (item 4)
51517462
8
Inp. module conductivity, conductive measurement, non-Ex
• FSLC1, insertion module, cond. conductivity, non-Ex
• Terminal strip (item 9)
51518005
8
Inp. module conductivity, inductive measurement, non-Ex
• FSLI1, insertion module, ind. conductivity, non-Ex
• Terminal strip (item 9)
51518006
9
Inp. module terminal set, conductivity, conductive measurement
51517489
9
Inp. module terminal set, conductivity, inductive measurement
51517490
10
Inp. module digital/Memosens, non-Ex
• FSDG1, insertion module, 1 channel, 1 SWU, non-Ex
• Terminal strip (item 11)
51518007
11
Inp. module terminal set, digital/Memosens
51517491
12
Inp. module pH/ORP/temperature, non-Ex
• FSPH1, insertion module, pH/mV, non-Ex
• Terminal strip (item 13)
51518004
13
Inp. module terminal set, pH/ORP/temperature
51517487
14
Threaded joints, G½
• 1 set of threaded joints
• Thread base plate
51517506
No Fig.
Terminal set, 5 pcs. each for all modules, 7 types
51517498
Endress+Hauser
Liquiline M CM42
Item
Troubleshooting - hardware
Kit CM42
Order number
No Fig.
10 set cable clamps+screws
• 10 EMC cable clamps D6 (0.24")
• 10 EMC cable clamps D4 (0.16")
• 10 cheese head screws M4x8 A2
51517499
No Fig.
Connection jack for external Historom/CDI
51517507
1)
8.3
When ordering, you must specify the serial number of the device for which you are ordering the spare part.
Return
If returning the transmitter, please send it cleaned to your sales center.
Use the original packaging when returning the device.
8.4
Disposal
This product contains electronic components. For this reason, it must be disposed of as
electronic waste.
Please observe local regulations.
Endress+Hauser
39
Accessories
9
!
Liquiline M CM42
Accessories
Note!
The CD-ROM supplied provides you with all the relevant information on an extensive range of
accessories, including descriptions, graphics and ordering information.
For example, it contains information on:
• Sensors
• Assemblies
• Quality buffers from Endress+Hauser
9.1
Installation kits
Post mount
• 1 Mounting plate
• 2 Securing screws
• Order No. 51518263
Panel installation kit for panel cutout 138x138 mm (5.43x5.43 inch)
• 1 Panel installation seal
• 2 Tensioning screws M6x150 mm
• 4 Hexagonal nuts M6, DIN934 A2
• 4 Spring washers, A2 DIN127, form B6
• 4 Washers A6.4, DIN125 A2
• Order No. 51518173
9.2
Weather protection
Weather protection cover
• Order No. 51517382
9.3
Communication
M12 connector
• Four-pole metal connector for mounting on transmitter
• For connecting to connection box or cable jack. Cable length 150 mm (5.91").
• Order No. 51502184
40
Endress+Hauser
Liquiline M CM42
10
Technical data
Technical data
10.1 Input
10.1.1 Measured variables
• pH value
• ORP
• Temperature
10.1.2 Measuring range
pH value
–2 to 16
ORP
–1500 to +1500 mV
Temperature
–25 to +150 °C (–10 to +300 °F)
10.1.3 Cable specification
Without SCS
Max. cable length 50 m (160 ft)
With SCS
Max. cable length 20 m (65 ft)
10.2 Output
10.2.1 Output signal
Profibus PA
10.2.2 Signal on alarm
Digital via fieldbus
10.2.3 Output distribution
pH value
Configurable, ΔpH > 0.5
ORP
Configurable, ΔU > 5 mV
Temperature
Adjustable, at least 2 °C (2 °F)
Endress+Hauser
41
Technical data
Liquiline M CM42
10.3 Electrical connection
10.3.1 Supply voltage and signal voltage
Profibus PA
9 to 32 V DC
Bus current consumption:
22 mA
10.3.2 Cable cross-section
Max. cable cross-section: 2.5 mm2 (i14 AWG), GND 4 mm2 (i12 AWG)
10.4 Performance characteristics
10.4.1 Reference temperature
25 °C (77 °F)
10.4.2 Measured value resolution
pH (glass electrodes and ISFET sensors)
0.01 pH
ORP
1 mV
Temperature
0.1 °C (0.1 °F)
SCS glass electrodes
Glass
Reference
Leakage current, ISFET sensors
42
100 kΩ
100 Ω
100 nA
Endress+Hauser
Liquiline M CM42
Technical data
10.4.3 Maximum measured error4)
pH (glass electrodes and ISFET sensors)
0.02 pH
ORP
1 mV
Temperature
NTC30k
Pt100 / Pt1000
0.5 K
0.25 K
SCS glass electrodes
Glass
Reference
Leakage current, ISFET sensors
200 MΩ
200 Ω
100 nA
10.4.4 Repeatability
pH
< 0.01
10.5 Environment
10.5.1 Ambient temperature range
–30 to 70 °C (–20 to 160 °F)
10.5.2 Ambient temperature limits
–30 to +80 °C (–20 to 175 °F)
10.5.3 Storage temperature
–40 to 80 °C (–40 to 175 °F)
10.5.4 Electromagnetic compatibility
Interference emission and interference immunity to EN 61326: 2004
10.5.5 Degree of protection
IP 67 (iNEMA 4X)
4)
In accordance with IEC 746-1 for nominal operating conditions
Endress+Hauser
43
Technical data
Liquiline M CM42
10.5.6 Relative humidity
10 to 95%, not condensing
10.6 Mechanical construction
10.6.1 Weight
1.5 kg (3.3 lbs)
10.6.2 Material
Housing
Housing seals
44
Polycarbonate
EPDM
Endress+Hauser
Liquiline M CM42
Index
A
Access authorization
Change . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
User roles. . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Access code
Factory setting . . . . . . . . . . . . . . . . . . . . . . . . 32
Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Communication . . . . . . . . . . . . . . . . . . . . . . . 40
Installation kits . . . . . . . . . . . . . . . . . . . . . . . . 40
Ambient temperature limits . . . . . . . . . . . . . . . . 43
Ambient temperature range . . . . . . . . . . . . . . . . 43
Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
B
Bus address . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
C
Cable cross-section . . . . . . . . . . . . . . . . . . . . . . . 42
Cable grounding . . . . . . . . . . . . . . . . . . . . . . . . . 18
Cable specification . . . . . . . . . . . . . . . . . . . . . . . 41
Cable terminals . . . . . . . . . . . . . . . . . . . . . . . . . 19
Calibrate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Check
Electrical connection . . . . . . . . . . . . . . . . . . . 27
Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Post-installation . . . . . . . . . . . . . . . . . . . . . . . 16
Commissioning. . . . . . . . . . . . . . . . . . . . . . . . 4, 28
Communication . . . . . . . . . . . . . . . . . . . . . . . . . 40
Configuring
Device address . . . . . . . . . . . . . . . . . . . . . . . . 28
Connecting ground. . . . . . . . . . . . . . . . . . . . . . . 18
Connection
Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Glass electrodes . . . . . . . . . . . . . . . . . . . . . . . 23
ISFET sensors . . . . . . . . . . . . . . . . . . . . . . . . . 24
Pfaudler electrodes . . . . . . . . . . . . . . . . . . . . . 25
With PM . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Without PM . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Connection data . . . . . . . . . . . . . . . . . . . . . . . . . 42
D
Degree of protection . . . . . . . . . . . . . . . . . . . . . . 43
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Device address . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Endress+Hauser
Device description. . . . . . . . . . . . . . . . . . . . . . . . 10
Device master files . . . . . . . . . . . . . . . . . . . . . . . 29
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Documentation
Methods of ordering . . . . . . . . . . . . . . . . . . . . . 8
E
Electrical connection . . . . . . . . . . . . . . . . . . . 27, 42
Power supply . . . . . . . . . . . . . . . . . . . . . . 19, 42
Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Signal output. . . . . . . . . . . . . . . . . . . . . . . 19, 42
Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Electrical technician . . . . . . . . . . . . . . . . . . . . . . 17
Electromagnetic compatibility . . . . . . . . . . . . . . . 43
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
F
Factory setting
Access code . . . . . . . . . . . . . . . . . . . . . . . 32–33
G
Glass electrodes. . . . . . . . . . . . . . . . . . . . . . . . . . 23
Grounding concept
Capacitive connection to ground . . . . . . . . . . . 20
Hard grounding. . . . . . . . . . . . . . . . . . . . . . . . 20
H
Housing
Closed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Open . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Housing grounding . . . . . . . . . . . . . . . . . . . . . . .
Housing opening . . . . . . . . . . . . . . . . . . . . . . . . .
10
10
18
17
I
Icons
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Identification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Serial number . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Incoming acceptance. . . . . . . . . . . . . . . . . . . . . . . 4
Input
Cable specification . . . . . . . . . . . . . . . . . . . . . 41
Measured variables . . . . . . . . . . . . . . . . . . . . . 41
Measuring range . . . . . . . . . . . . . . . . . . . . . . . 41
Liquiline M CM42
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 11
Installation instructions
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Panel mounting . . . . . . . . . . . . . . . . . . . . . . . 15
Wall or field mounting . . . . . . . . . . . . . . . . . . 14
Installation kits . . . . . . . . . . . . . . . . . . . . . . . . . . 40
ISFET sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
35
44
43
42
41
41
44
44
13
N
Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
O
Opening the housing. . . . . . . . . . . . . . . . . . . . . . 17
Operation
Safety instructions. . . . . . . . . . . . . . . . . . . . . . . 4
Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . 5
Output
Output distribution . . . . . . . . . . . . . . . . . . . . . 41
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . 41
Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . 41
P
Performance characteristics
Maximum measured error . . . . . . . . . . . . . . . 43
Measured value resolution . . . . . . . . . . . . . . . 42
Reference temperature . . . . . . . . . . . . . . . . . . 42
Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Pfaudler electrodes . . . . . . . . . . . . . . . . . . . . . . . 25
PLC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Post-installation
Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Power supply . . . . . . . . . . . . . . . . . . . . . . . . 19, 42
Product structure . . . . . . . . . . . . . . . . . . . . . . . . . 7
Q
Quick Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
46
Reference temperature. . . . . . . . . . . . . . . . . . . . 42
Relative humidity . . . . . . . . . . . . . . . . . . . . . . . 44
Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Replacing modules. . . . . . . . . . . . . . . . . . . . . . . 36
Return. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5, 39
S
M
Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Maximum measured error . . . . . . . . . . . . . . . . .
Measured value resolution . . . . . . . . . . . . . . . . .
Measured variables . . . . . . . . . . . . . . . . . . . . . . .
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . .
Mechanical construction
Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Mounting plate . . . . . . . . . . . . . . . . . . . . . . . . . .
R
Safety conventions and icons . . . . . . . . . . . . . . . . 5
Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . 4
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . 8
Sensor
Connection with PM . . . . . . . . . . . . . . . . . . . 23
Connection without PM . . . . . . . . . . . . . . . . 26
Electrical connection . . . . . . . . . . . . . . . . . . . 22
Glass electrodes. . . . . . . . . . . . . . . . . . . . . . . 23
ISFET sensors . . . . . . . . . . . . . . . . . . . . . . . . 24
Pfaudler electrodes . . . . . . . . . . . . . . . . . . . . 25
Serial number . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Signal output . . . . . . . . . . . . . . . . . . . . . . . . 19, 42
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Storage temperature . . . . . . . . . . . . . . . . . . . . . 43
Symbols
Document . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
System setup . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
T
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Input. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Mechanical construction . . . . . . . . . . . . . . . . 44
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Performance characteristics . . . . . . . . . . . . . . 42
Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . 36
Type files. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Types of calibration . . . . . . . . . . . . . . . . . . . . . . 34
U
Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
User administration . . . . . . . . . . . . . . . . . . . . . . 32
User groups . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
User roles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
W
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Weather protection . . . . . . . . . . . . . . . . . . . . . . 40
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Wiring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Liquiline M CM42
Endress+Hauser
www.endress.com/worldwide
BA381C/07/en/POD
Printed in Germany / FM+SGML 6.0