Emerson Clarity II™ Turbidimeter Turbidity Measurement System Instruction Manual
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Instruction Manual
PN 51T1056/rev.D
August 2014
Clarity II™ Turbidimeter
Turbidity Measurement System
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Your instrument purchase from Rosemount
Analytical, Inc. is one of the finest available for your
particular application. These instruments have been
designed, and tested to meet many national and
international standards. Experience indicates that its
performance is directly related to the quality of the
installation and knowledge of the user in operating
and maintaining the instrument. To ensure their con
tinued operation to the design specifications, per
sonnel should read this manual thoroughly before
proceeding with installation, commissioning, opera
tion, and maintenance of this instrument. If this
equipment is used in a manner not specified by the
manufacturer, the protection provided by it against
hazards may be impaired.
• Failure to follow the proper instructions may
cause any one of the following situations to
occur: Loss of life; personal injury; property dam
age; damage to this instrument; and warranty
invalidation.
• Ensure that you have received the correct model
and options from your purchase order. Verify that
this manual covers your model and options. If
not, call 18008548257 or 9497578500 to
request correct manual.
• For clarification of instructions, contact your
Rosemount representative.
• Follow all warnings, cautions, and instructions
marked on and supplied with the product.
• Use only qualified personnel to install, operate,
update, program and maintain the product.
• Educate your personnel in the proper installation,
operation, and maintenance of the product.
• Install equipment as specified in the Installation
section of this manual. Follow appropriate local
and national codes. Only connect the product to
electrical and pressure sources specified in this
manual.
• Use only factory documented components for
repair. Tampering or unauthorized substitution of
parts and procedures can affect the performance
and cause unsafe operation of your process.
• All equipment doors must be closed and protec
tive covers must be in place unless qualified per
sonnel are performing maintenance.
Emerson Process Management
2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 7578500
Fax: (949) 4747250
http://www.rosemountanalytical.com
© Rosemount Analytical Inc. 2014
WARNING
RISK OF ELECTRICAL SHOCK
Equipment protected throughout by double insulation.
• Installation and servicing of this product may expose personel
to dangerous voltages.
• Main power wired to separate power source must be
disconnected before servicing.
• Do not operate or energize instrument with case open!
• Signal wiring connected in this box must be rated at least
240 V.
• Nonmetallic cable strain reliefs do not provide grounding
between conduit connections! Use grounding type bushings
and jumper wires.
• Unused cable conduit entries must be securely sealed by
nonflammable closures to provide enclosure integrity in
compliance with personal safety and environmental protection
requirements. Unused conduit openings must be sealed with
NEMA 4X or IP65 conduit plugs to maintain the ingress
protection rating (NEMA 4X).
• Electrical installation must be in accordance with the National
Electrical Code (ANSI/NFPA70) and/or any other applicable
national or local codes.
• Operate only with front panel fastened and in place.
• Proper use and configuration is the responsibility of the user.
CAUTION
This product generates, uses, and can radiate radio frequency
energy and thus can cause radio communication interference.
Improper installation, or operation, may increase such interfer
ence. As temporarily permitted by regulation, this unit has not
been tested for compliance within the limits of Class A comput
ing devices, pursuant to Subpart J of Part 15, of FCC Rules,
which are designed to provide reasonable protection against
such interference. Operation of this equipment in a residential
area may cause interference, in which case the user at his own
expense, will be required to take whatever measures may be
required to correct the interference.
CAUTION
This product is not intended for use in the light industrial,
residential or commercial environments per the instru
ment’s certification to EN500812.
QUICK START GUIDE
FOR CLARITY II TURBIDIMETER
1. Refer to Section 2.0 for installation instructions.
2. The sensor cable is prewired to a plug that inserts into a receiving socket in the analyzer. The cable also passes
through a strain relief fitting. To install the cable…
a. Remove the wrenching nut from the strain relief fitting.
b. Insert the plug through the hole in the bottom of the enclosure nearest the sensor socket. Seat the fitting in the
hole.
c.
Slide the wrenching nut over the plug and screw it onto the fitting.
d. Loosen the cable nut so the cable slides easily.
e. Insert the plug into the appropriate receptacle on the circuit board.
f.
Adjust the cable slack in the enclosure and tighten the cable nut. For wall/pipe mounting, be sure to leave suffi
cient cable in the enclosure to avoid stress on the cable and connections.
g. Plug the cable into the back of the sensor.
h. Place the sensor in either the measuring chamber or the calibration cup. The sensor must be in a dark place
when power is first applied to the analyzer.
3. Make power, alarm, and output connections as shown in section 3.0 wiring.
4. Once connections are secured and verified, apply power to the analyzer.
5. When the analyzer is powered up for the first time Quick Start screens appear. Follow the Quick Start Guide to
enable live readings.
a. A blinking field shows the position of the cursor.
b. Use the t or u key to move the cursor left or right. Use the p or q key to increase or decrease the value of a
digit. Use the p or q key to move the decimal point.
c.
Press ENTER to store a setting. Press EXIT to leave without storing changes. Pressing EXIT also returns the
display to the language selection screen.
IMPORTANT NOTE:
When using EPA/incandescent sensors (PN 801080002EPA):
n
DO NOT power up the instrument without the sensor connected
n
DO NOT disconnect and reconnect a sensor while an analyzer is powered
If this is inconvenient or cannot be avoided:
1. Cycle power to the instrument after connecting the sensor or..
2. Perform a Slope Calibration or Standard Calibration routine after connecting the sensor. Following these guide
lines will extend the life of the incandescent lamp and avoid premature warnings and faults due to reduced
lamp life.
CONTINUED ON THE FOLLOWING PAGE
Figure A. QUICK START GUIDE, Model 1056
QUICK START GUIDE
Figure B. MODEL 1056 MENU TREE
QUICK REFERENCE GUIDE
About This Document
This manual contains instructions for installation and operation of the Clarity II Model
T1056 Turbidimeter.
The following list provides notes concerning all revisions of this document.
Rev. Level
Date
Notes
A
12/07
This is the initiallaunch version.
B
2/09
Update Specifications
C
12/09
Update DNV logo, company name
D
08/14
Change turbidity specifications, add CSA Nonincendive field
wiring installation drawings
MODEL CLARITY II
TABLE OF CONTENTS
MODEL CLARITY II TURBIDIMETER
TABLE OF CONTENTS
Section Title
Page
1.0
1.1
1.2
DESCRIPTION AND SPECIFICATIONS ..............................................................................
Features and Applications ......................................................................................................
Specifications .........................................................................................................................
1
1
2
2.0
2.1
2.2
2.3
2.4
2.5
INSTALLATION......................................................................................................................
Unpacking and Inspection ......................................................................................................
Installation .............................................................................................................................
Installation — Debubbler Assembly........................................................................................
Installation — Sensor .............................................................................................................
Sample Point ..........................................................................................................................
5
5
5
8
10
10
3.0
3.1
3.2
3.3
3.4
WIRING ..................................................................................................................................
General...................................................................................................................................
Preparing Conduit Openings .................................................................................................
Preparing Sensor Cable .........................................................................................................
Power, Output, and Sensor Connections ...............................................................................
13
13
13
14
14
4.0
4.1
4.2
4.3
4.4
DISPLAY AND OPERATION .................................................................................................
User Interface.........................................................................................................................
Instrument Keypad .................................................................................................................
Main Display...........................................................................................................................
Menu System .........................................................................................................................
19
19
19
20
21
5.0
5.1
5.2
5.3
5.4
5.5
5.6
5.7
5.8
6.0
6.1
6.2
6.3
6.4
PROGRAMMING THE ANALYZER .......................................................................................
General...................................................................................................................................
Changing StartUp Settings .....................................................................................................
Configuring and Ranging the Current Outputs .......................................................................
Setting a Security Code..........................................................................................................
Security Access ......................................................................................................................
Using Hold..............................................................................................................................
Resetting Factory Default Settings.........................................................................................
Programming Alarm Relays....................................................................................................
PROGRAMMING TURBIDITY ...............................................................................................
Programming Measurements Introduction ...........................................................................
Turbidity Measurement Programming ....................................................................................
Choosing Turbidity or Total Suspended solids .......................................................................
Entering a Turbidity to TSS Conversion Equation ..................................................................
23
23
23
23
25
26
26
27
28
31
31
32
35
38
7.0
7.1
7.2
8.0
8.1
8.2
8.3
8.4
CALIBRATION .......................................................................................................................
Calibration Introduction ..........................................................................................................
Turbidity Calibration................................................................................................................
MAINTENANCE ....................................................................................................................
Model 1056.............................................................................................................................
Sensor ....................................................................................................................................
Debubbler and Measuring Chamber ......................................................................................
List of Replacement Parts ......................................................................................................
41
41
41
45
45
46
47
48
9.0
9.1
9.2
9.3
9.4
TROUBLESHOOTING...........................................................................................................
Overview ................................................................................................................................
Troubleshooting Using Fault Codes .......................................................................................
Troubleshooting Calibration Problems ...................................................................................
Troubleshooting Other Problems............................................................................................
49
49
45
47
52
10.0
RETURN OF MATERIAL ......................................................................................................
55
10.1
General...................................................................................................................................
56
10.2
Warranty Repair .....................................................................................................................
57
10.3
NONWarranty Repair ............................................................................................................
57
i
MODEL CLARITY II
TABLE OF CONTENTS
LIST OF FIGURES
Number Title
Page
21
Panel Mount Dimensions .........................................................................................
6
22
Pipe and Wall Mount Dimensions.............................................................................
7
23
Debubbler and Flow Chamber .................................................................................
9
24
Sensor
....................................................................................................
10
25
Sampling for Turbidity...............................................................................................
10
26
CSA Nonincendive Field Wiring Installation for Turbidity, page 1............................
11
27
CSA Nonincendive Field Wiring Installation for Turbidity, page 2............................
12
31
24VDC Power Supply...............................................................................................
14
32
Switching AC Power Supply .....................................................................................
14
33
Current Output Wiring ..............................................................................................
15
34
Alarm Relay Wiring...................................................................................................
15
35
Turbidity Signal Board ..............................................................................................
16
36
Power Wiring for Model 1056 85265 VAC Power Supply .......................................
17
37
Power Wiring for Model 1056 24VDC Power Supply ...............................................
17
38
Output Wiring for Model 1056 Main PCB .................................................................
18
41
Formatting the Main Display.....................................................................................
22
52
Configuring and Ranging the Current Outputs ........................................................
24
53
Setting a Security Code............................................................................................
25
54
Using Hold................................................................................................................
26
55
Resetting Factory Default Settings...........................................................................
27
61
Converting Turbidity to TSS .....................................................................................
33
64
Turbidity Sensor — General .....................................................................................
35
65
Turbidity Sensor — EPA 108.1 .................................................................................
35
66
Turbidity Sensor — ISO 7027...................................................................................
36
62
Converting Turbidity to TSS .....................................................................................
38
62
Lowest Turbidity (TSS) .............................................................................................
38
67
Configure Turbidity Measurement ............................................................................
40
78
Calibrate Turbidity ....................................................................................................
44
83
Replacing the Lamp/LED Board...............................................................................
46
84
Molded Buffer Assembly...........................................................................................
48
LIST OF TABLES
Number Title
Page
21
Approximate Debubbler Pressure as a Function of Flow ........................................
8
611
Turbidity Measurement Programming ......................................................................
32
712
Turbidity Calibration Routines .....................................................................................
41
81
Replacement Parts for Model 1506 ..........................................................................
45
APPENDIX
APPENDIX ..............................................................................................................
ii
56
MODEL CLARITY II
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
SECTION 1.0.
DESCRIPTION AND SPECIFICATIONS
• COMPLETE SYSTEM includes single or dual input analyzer, sensor(s), and debubbler
assembly
• CHOOSE U.S. EPA METHOD 180.1 or ISO METHOD 7027 compliant sensors
• RESOLUTION 0.001 NTU
• FULL FEATURED ANALYZER with fully scalable analog outputs and fully programmable
alarms with interval timers
• INTUITIVE, USERFRIENDLY MENU in seven languages makes setup and calibration
easy
Clarity II is a trademark of Emerson Process Management.
1.1 FEATURES AND APPLICATIONS
The Clarity II turbidimeter is intended for the determi
nation of turbidity in water. Low stray light, high stability,
efficient bubble rejection, and a display resolution of
0.001 NTU make Clarity II ideal for monitoring the
turbidity of filtered drinking water. The Clarity II
turbidimeter can be used in applications other than
drinking water treatment. Examples are monitoring
wastewater discharges, condensate returns, and clarifiers.
The Clarity II turbidimeter incorporates the popular
and easy to use Model 1056 analyzer. Menu flows
and prompts are so intuitive that a manual is practically
not needed. Analog outputs are fully scalable. Alarms
are fully programmable for high/low logic and dead
band. To simplify programming, the analyzer automati
cally detects whether an EPA 180.1 or ISO 7027
sensor is being used.
Both USEPA 180.1 and ISO 7027compliant sensors
are available. USEPA 180.1 sensors use a visible light
source. ISO 7027 sensors use a near infrared LED.
For regulatory monitoring in the United States, USEPA
180.1 sensors must be used. Regulatory agencies in
other countries may have different requirements.
Clarity II is available in an optional configuration in
which the analyzer, sensor(s), and debubbling flow
cell(s) are mounted on a single back plate. The sensor
cables are prewired to the analyzer, so setup is
exceptionally fast and easy. All the user does is mount
the unit on a wall, bring in power and sample, and pro
vide a drain. To order this option, consult the factory.
The Clarity II turbidimeter consists of an analyzer,
which accepts either one or two sensors, the sensors
themselves, and a debubbler/measuring chamber and
cable for each sensor. The cable plugs into the sensor
and the analyzer, making setup fast and easy. Sensors
can be located as far as 50 ft (15.2 m) away from the
analyzer.
1
MODEL CLARITY II
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
SPECIFICATIONS General
Enclosure: Polycarbonate. Type 4X/CSA 4 (IP65).
Dimensions: Overall 155 x 155 x 131mm (6.10 x 6.10
x 5.15 in.). Cutout: 1/2 DIN 139mm x 139mm (5.45 x
5.45 in.)
Power: Code 02: 20 to 30 VDC. 15 W.
Code 03: 85 to 265 VAC, 47.5 to 65.0 Hz, switching.
15 W.
Note: Code 02 and 03 power supplies include 4 pro
grammable relays
Equipment protected by double insulation
RFI/EMI: EN61326
LVD:
EN610101
Alarms relays*: Four alarm relays for process meas
urement(s) or temperature. Any relay can be config
ured as a fault alarm instead of a process alarm. Each
relay can be configured independently and each can
be programmed with interval timer settings.
Conduit Openings: Accepts 1/2” or PG13.5 conduit
fittings
Display: Monochromatic graphic liquid crystal display.
128 x 96 pixel display resolution. Backlit. Active
display area: 58 x 78mm (2.3 x 3.0 in.).
Ambient Temperature and Humidity: 0 to 55°C
(32 to 131°F). Turbidity only: 0 to 50°C (32 to
122°F), RH 5 to 95% (noncondensing)
Storage Temperature Effect: 20 to 60ºC (4 to 140°F)
Hazardous Location Approvals
Options for CSA: 02, 03, 20, 21, 22, 24, 25, 26, 27, 30,
31, 32, 34, 35, 36, 37, 38, AN, and HT.
Class I, Division 2, Groups A, B, C, & D
Class Il, Division 2, Groups E, F, & G
Class Ill T4A Tamb= 50° C
Evaluated to the ANSI/UL Standards. The ‘C’ and ‘US’ indi
cators adjacent to the CSA Mark signify that the product has
been evaluated to the applicable CSA and ANSI/UL
Standards, for use in Canada and the U.S. respectively
POLLUTION DEGREE 2: Normally only nonconductive
pollution occurs. Occasionally, however, a temporary
conductivity caused by condensation must be expected.
Altitude: for use up to 2000 meter (6562 ft.)
2
Relays: Form C, SPDT, epoxy sealed
Maximum Relay Current
Resistive
28 VDC
115 VAC
230 VAC
5.0 A
5.0 A
5.0 A
Inductive load: 1/8 HP motor (max.), 40 VAC
CAUTION
RISK OF ELECTRICAL SHOCK
*Relays only available with 02 power supply (20 30 VDC) or 03
switching power supply (85 265 VAC)
WARNING
WARNING
Exposure to some chemicals may degrade the
sealing properties used in the following devices:
Zettler Relays (K1K4) PN AZ81CH12DSEA
Inputs: One or two isolated sensor inputs
Outputs: Two 420 mA or 020 mA isolated current out
puts. Fully scalable. Max Load: 550 Ohm. Output 1
has superimposed HART signal (configurations
10560X2X3XHT only)
Current Output Accuracy: ±0.05 mA @ 25 ºC
Terminal Connections Rating: Power connector
(3leads): 2412 AWG wire size. Signal board ter
minal blocks: 2616 AWG wire size. Current output
connectors (2leads): 2416 AWG wire size. Alarm
relay terminal blocks: 2412 AWG wire size
(02 24 VDC power supply and 03 85265VAC
power supply)
MODEL CLARITY II
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
SPECIFICATIONS — SENSOR
Method: EPA 180.1 or ISO 7027 (using 860 nm LED
source). Must be specified when ordering.
Incandescent lamp life: two years
LED life: five years
Wetted materials: Delrin1, glass, EPDM
Accuracy after calibration at 20.0 NTU:
0 1 NTU: ±2% of reading or ±0.015 NTU,
whichever is greater.
0 20 NTU: ±2% of reading
Note: Turbidity values of 2200 NTU can be meas
ured, but frequent cleaning may be required to main
tain turbidity measurements.
Cable: 20 ft (6.1 m) or 50 ft (15.2 m). Maximum 50 ft
(15.2 m). Connector is IP65.
Maximum Pressure: 30 psig (308 kPa abs)
Temperature: 40 95°F (5 35°C)
Sensor body rating: IP65 when cable is connected
SPECIFICATIONS — DEBUBBLER AND
FLOW CHAMBER
Dimensions: 18.1 in. x 4.1 in. diam. (460 mm x 104 mm
diam.) (approx.)
Wetted materials: ABS, EPDM, Delrin1, polypropylene,
nylon
Inlet: compression fitting accepts 1/4 in. OD tubing; fit
ting can be removed to provide 1/4 in. FNPT
Drain: barbed fitting accepts 3/8 in. ID tubing; fitting can
be removed to provide 1/4 in. FNPT. Must drain to
atmosphere.
Sample temperature: 40 95°F (5 35°C)
Minimum inlet pressure : 3.5 psig (125 kPa abs). 3.5
psig will provide about 250 mL/min sample flow.
Maximum inlet pressure: 30 psig (308 kPa abs). Do
not block drain tube.
Recommended sample flow: 250 750 mL/min
Response Time: The table shows the time in minutes to
percent of final value following a step change in turbid
ity.
1Delrin is a registered trademark of DuPont Performance
Elastomers.
SPECIFICATIONS — MISCELLANEOUS
Weight/shipping weight:
Sensor: 1 lb/2 lb (0.5 kg/1.0 kg)
Analyzer: 2 lb/3 lb (1.0 kg/1.5 kg)
Debubbler: 3 lb/4 lb (1.5 kg/2.0 kg)
(rounded to the nearest lb or 0.5 kg
3
This page left blank intentionally.
4
MODEL CLARITY II
SECTION 2.0
INSTALLATION
SECTION 2.0.
INSTALLATION
2.1 UNPACKING AND INSPECTION
2.2 INSTALLATION
2.1 UNPACKING AND INSPECTION
The Clarity II Turbidimeter is a complete system for the determination of turbidity in drinking water. The system
consists of the analyzer, sensor(s), cable(s), and flow chamber/debubbler(s). Consult the table to verify that you
have received the parts for the option you ordered.
Item
Single Input Turbidity Analyzer
Model/part number
1056032738AN
Dual Input Turbidity Analyzer
1056032737AN
Single Input Turbidity Analyzer with HART
1056032738HT
Dual Input Turbidity Analyzer with HART
1056032737HT
SensorEPA standard
801080002EPA
SensorISO standard
801080003ISO
Cable3 ft (0.9 m)
2413800
Cable20 ft (6.1 m)
2409700
Cable50 ft (15.2 m)
2409800
Calibration cup
2410100
Molded chamber/debubbler
2417000
(1)
The analyzer model number is printed on a label attached to the side of the instrument.
2.2 INSTALLATION
2.2.1 General Information
1. Although the analyzer is suitable for outdoor use, do not install it in direct sunlight or in areas of extreme
temperatures.
2. Install the analyzer in an area where vibration and electromagnetic and radio frequency interference are
minimized or absent.
3. Keep the analyzer and sensor wiring at least one foot from high voltage conductors. Be sure there is easy
access to the analyzer.
4. The analyzer is suitable for panel, pipe, or surface mounting. Refer to the table below.
Type of Mounting
Figure
Panel
21
Wall and Pipe
22
WARNING
WARNING
RISK OF ELECTRICAL SHOCK
Electrical installation must be in accordance with the
National Electrical Code (ANSI/NFPA70) and/or any
other applicable national or local codes.
5
FIGURE 21 PANEL MOUNTING DIMENSIONS
MILLIMETER
INCH
17.13
1.1
126.4
5.0
101.6
4.00
154.9
6.1
154.9
6.1
Front View
Side View
( 126.4
5.0 )
76.2
3.0
41.4
1.6
Bottom View
152.73
6.0
Note: Panel mounting seal integrity (4/4X) for outdoor applications is the responsibility of the end user.
6
FIGURE 22 PIPE AND WALL MOUNTING DIMENSIONS
(Mounting bracket PN:2382000)
MILLIMETER
INCH
154.9
6.1
Wall / Surface Mount
232
9.1
102
4.0
33.5
1.3
130
5.1
187
7.4
154.9
6.1
165
6.5
Side View
Front View
Pipe Mount
232
9.1
Bottom View
33.5
1.3
130
5.1
80.01
3.2
165
6.5
45.21
1.8
108.9
4.3
Side View
71.37
2.8
The front panel is hinged at the bottom. The panel swings down for easy access to the wiring locations.
7
MODEL CLARITY II
SECTION 2.0
INSTALLATION
2.3 INSTALLATION — DEBUBBLER ASSEMBLY
See Figure 23 for installation.
Connect the sample line to the inlet fitting. The fitting accepts 1/4inch OD tubing. See Section 2.6 for recom
mended installation of the sample port.
Attach a piece of 3/8 inch ID soft tubing to the drain fitting. The debubbler must drain to atmosphere.
NOTE
During operation, the debubbler is under pressure. A 0.040 inch (1 mm) orifice in the outlet pro
vides the pressure. Back pressure helps prevent outgassing, which can lead to bubbles accu
mulating on the sensor face resulting in erroneous readings. DO NOT EXCEED 30 psig (308
kPa abs) inlet pressure.
WARNING
Before removing the sensor, be absolutely certain
that the process pressure is reduced to 0 psig and
the process temperature is lowered to a safe level!
The amount of pressure in the debubbler can be estimated from the flow rate. See Table 21.
To control and monitor sample flow, a valved rotameter with fittings is available (PN 2410300). Attach the
rotameter to the debubbler outlet. The rotameter can also be used to increase back pressure on the debubbler if
additional pressure is needed to prevent outgassing.
TABLE 21. Approximate debubbler pressure
as a function of flow (0.040 inch outlet orifice)
8
gph
psig
mL/min
kPa abs
2
1
100
110
4
3
200
120
6
8
300
140
8
14
400
160
10
21
500
190
11
26
600
240
12
31
700
280
—
—
800
340
MODEL CLARITY II
SECTION 2.0
INSTALLATION
INCH
MILLIMETER
FIGURE 23. Debubbler and Flow Chamber
9
MODEL CLARITY II
SECTION 2.0
INSTALLATION
2.4 INSTALLATION — SENSOR
Unscrew the nut on the side of the debubbler. Insert the sensor in the mouth of the measuring chamber. Be sure
the pin on the debubbler lines up with the hole in the sensor. Replace the nut. Remove the protective cap from
the sensor and screw the cable onto the receptacle. The plug and receptacle are keyed for proper alignment.
The sensor is rated to IP65 when properly connected to the cable. To prevent possible water damage to the
connector contacts, be sure the cable receptacle and the connector on the back of the sensor are dry when
connecting or disconnecting the cable.
INCH
MILLIMETER
DWG. NO.
40T105501
REV.
A
FIGURE 24. Sensor
2.5 SAMPLE POINT
Locate the sample tap to minimize pickup of sediment or air. See Figure 25. If possible, install a sampling port
that extends one or two inches (25 50 mm) into the pipe. Use ¼ inch OD rigid plastic tubing. Avoid soft plastic
tubing if possible. To reduce sample lag time, install the debubbler and flow chamber as close to the sample tap
as possible.
FIGURE 25. Sampling for Turbidity
10
11
3
METAL CONDUIT
METAL CONDUIT
ANALOG OUTPUT
GROUND CONNECTION MAY BE MADE IN HAZARDOUS AREA.
SEAL REQUIRED AT EACH CONDUIT ENTRANCE.
INSTALLATION MUST CONFORM TO THE CEC.
3
2.
1.
FINISH
WARNING
5
6
B. JOHNSON
THIS FILE CREATED USING
SOLID EDGE
ENG APVD J. FLOCK
CHECKED J. FLOCK
DRAWN
APPROVALS
6
1/5/05
1/5/05
Emerson
D3
REV
REV
REV
REV
REV
REV
C
REVISIONS NOT PERMITTED
W/O AGENCY APPROVAL
CSA
THIS DOCUMENT IS
CERTIFIED BY
1400314
SCALE: NONE WEIGHT:
C
SIZE DWG NO
SHEET 1 OF 1
REV
D
NON INCENDIVE FIELD
WIRING INSTALLATION (CSA)
1056-27/37
ANALYTICAL
ROSEMOUNT
CLARITY II
TURBIDITY
SENSOR #1
CLARITY II
TURBIDITY
SENSOR #2
(OPTIONAL)
BJ
PROCESS MANAGEMENT
12/8/04 TITLE
DATE
UNCLASSIFIED
AREA
SENSOR CABLE
IS SHIELDED
5
FOR USE WITH NON-FLAMMABLE
PROCESS MEDIA ONLY
!
SEE ECO
SENSOR CABLE
IS SHIELDED
0000
DIMENSIONS ARE IN INCHES
REMOVE BURRS & SHARP EDGES
MACHINE FILLET RADII .020 MAX
NOMINAL SURFACE FINISH: 125
ANGLES ± 1/2°.
.XX ± .03 .XXX ± .010
MATERIAL
Figure 26
DURING INSTALLATION, LEAVE MAXIMUM AMOUNT OF JACKET INSULATION POSSIBLE ON N.I. FIELD
WIRING WITHIN INSTRUMENT ENCLOSURE. AFTER TERMINATION, WRAP N.I. FIELD WIRING WITHIN
ENCLOSURE WITH MYLAR TAPE, TO ENSURE ADEQUATE DOUBLE INSULATION REMAINS.
4
NOTES: UNLESS OTHERWISE SPECIFIED
MAX CABLE LENGTH IS 50 FEET.
CLASS I, DIV. 2, GPS A-D, 0° - 50 ° C
CLASS II, III, DIV 2, GPS E-G
1056-27/37
HAZARDOUS AREA
ALARM
WIRING (VAC)
(OPTIONAL)
5
UNCLASSIFIED
AREA
D2
POWER SUPPLY
METAL CONDUIT
D1
D
12
Figure 27
MODEL CLARITY II
SECTION 3.0
WIRING
SECTION 3.0.
WIRING
3.1
3.2
3.3
3.4
GENERAL
PREPARING CONDUIT OPENINGS
PREPARING SENSOR CABLE
POWER, OUTPUT, AND SENSOR
CONNECTIONS
3.1 GENERAL
The 1056 is easy to wire. It includes removable connectors and slideout signal input boards.
3.1.1. Removable connectors and signal input boards
Model 1056 uses removable signal input boards and communication boards for ease of wiring and instal
lation. Each of the signal input boards can be partially or completely removed from the enclosure for wiring.
The Model 1056 has three slots for placement of up to two signal input boards and one communication
board.
Slot 1Left
Slot 2 – Center
Slot 3 – Right
Comm. board
Input Board 1
Input Board 2
3.1.2. Signal Input boards
Slots 2 and 3 are for signal input measurement boards. Wire the sensor leads to the measurement board
following the lead locations marked on the board. After wiring the sensor leads to the signal board, carefully slide
the wired board fully into the enclosure slot and take up the excess sensor cable through the cable gland. Tighten
the cable gland nut to secure the cable and ensure a sealed enclosure.
3.1.3. Digital Communication boards
HART and Profibus DP communication boards will be available in the future as options for Model 1056 digital
communication with a host. The HART board supports Bell 202 digital communications over an analog
420mA current output. Profibus DP is an open communications protocol which operates over a dedicated
digital line to the host.
3.1.4 Alarm relays
Four alarm relays are supplied with the switching power supply (85 to 265VAC, 03 order code) and the 24VDC
power supply (2030VDC, 02 order code). All relays can be used for process measurement(s) or temperature.
Any relay can be configured as a fault alarm instead of a process alarm. Each relay can be configured
independently and each can be programmed as an interval timer, typically used to activate pumps or control
valves. As process alarms, alarm logic (high or low activation or USP*) and deadband are userprogrammable.
Customerdefined failsafe operation is supported as a programmable menu function to allow all relays to be
energized or notenergized as a default condition upon powering the analyzer.
The USP* alarm can be programmed to activate when the conductivity is within a userselectable
percentage of the limit. USP alarming is available only when a contacting conductivity measurement board is
installed.
3.2 PREPARING CONDUIT OPENINGS
There are six conduit openings in all configurations of Model 1056. (Note that four of the openings will be fitted
with plugs upon shipment.)
Conduit openings accept 1/2inch conduit fittings or PG13.5 cable glands. To keep the case watertight, block
unused openings with Type 4X or IP65 conduit plugs.
NOTE: Use watertight fittings and hubs that comply with your requirements. Connect the conduit hub to the
conduit before attaching the fitting to the analyzer.
13
MODEL CLARITY II
SECTION 3.0
WIRING
3.3 PREPARING SENSOR CABLE
The Model 1056 is intended for use with all Rosemount Analytical sensors. Refer to the sensor installation instructions
for details on preparing sensor cables.
3.4 POWER, OUTPUT, AND SENSOR CONNECTIONS
3.4.1 Power wiring
Two Power Supplies are offered for Model T1056:
a. 24VDC (20 – 30V) Power Supply (02 ordering code)
b. 85 – 265 VAC Switching Power Supply (03 ordering code)
AC mains (115 or 230V) leads and 24VDC leads are wired to the Power Supply board which is mounted vertically
on the left side of the main enclosure cavity. Each lead location is clearly marked on the Power Supply board.
Wire the power leads to the Power Supply board using the lead markings on the board.
24VDC Power Supply (02 ordering code) is shown below:
This power supply automatically
detects DC power and accepts 20VDC
to 30VDC inputs.
Four programmable alarm relays are
included.
Figure 31
Switching AC Power Supply (03 ordering code) is shown below:
This power supply automatically
detects AC line conditions and switches
to the proper line voltage and line
frequency.
Four programmable alarm relays are
included.
Figure 32
14
MODEL CLARITY II
SECTION 3.0
WIRING
3.4.2 Current Output wiring
All instruments are shipped with two 420mA current
outputs. Wiring locations for the outputs are on the
Main board which is mounted on the hinged door of the
instrument. Wire the out put leads to the correct posi
tion on the Main board using the lead markings (+/positive,
/negative) on the board. Male mating connectors are
provided with each unit.
Keep sensor and output signal wiring separate from
power wiring. Do not run sensor and power wiring in the
same conduit or close together in a cable tray.
For best EMI/RFI protection use shielded output signal
cable enclosed in an earthgrounded metal conduit.
Connect the shield to earth ground. AC wiring should
be 14 gauge or greater. Provide a switch or breaker to
disconnect the analyzer from the main power supply.
Install the switch or breaker near the analyzer and label
it as the disconnecting device for the analyzer.
3.4.3
Figure 3.3
Alarm relay wiring
Four alarm relays are supplied with the switching power supply (85 to 265VAC, 03 order code) and the 24VDC
power supply (2030VDC, 02 order code). Wire the relay leads on each of the independent relays to the correct
position on the power supply board using the printed lead markings (NO/Normally Open, NC/Normally Closed, or
Com/Common) on the board. See Fig 33.
NO1
COM1
RELAY 1
NC1
NO2
COM2
RELAY 2
NC2
NO3
COM3
RELAY 3
NC3
NO4
COM4
RELAY 4
NC4
Figure 34 Alarm Relay Wiring for Model 1056 Switching Power Supply (03 Order Code)
3.4.4 Sensor wiring to signal boards
Plug the preterminated sensor cable connector directly into the turbidity signal board mating connector.
WARNING
WARNING
RISK OF ELECTRICAL SHOCK
Electrical installation must be in accordance with the
National Electrical Code (ANSI/NFPA70) and/or any
other applicable national or local codes.
15
MODEL CLARITY II
SECTION 3.0
WIRING
3.2.2 Sensor
The sensor cable is prewired to a plug that inserts into a receiving socket on the signal board. See Figures 31.
The cable also passes through a strain relief fitting. To install the cable…
1. Remove the wrenching nut from the strain relief fitting.
2. Insert the plug through the hole in the bottom of the enclosure nearest the sensor socket. Seat the fitting in the
hole.
3. Slide the wrenching nut over the cable plug and screw it onto the fitting.
4. Loosen the cable nut so the cable slides easily.
5. Insert the plug into the appropriate receptacle. To remove the plug, squeeze the release clip and pull straight
out.
6. Adjust the cable slack in the enclosure and tighten the cable nut. Be sure to allow sufficient slack to avoid plac
ing stress on the cable and connections.
7. Plug the cable into the back of the sensor. The sensor is rated to IP65 when properly connected to the cable.
To prevent possible water damage to the connector contacts, be sure the cable receptacle and the connec
tor on the back of the sensor are dry when connecting or disconnecting the cable.
8. Place the sensor in either the measuring chamber or the calibration cup. The sensor must be in a dark place
when power is first appled to the analyzer.
Note: If “S1 Warning” appears, check sensor cable connection and confirm sample water flow at debubbler
drain outlet.
Figure 35 Turbidity signal board with plugin Sensor connection
IMPORTANT NOTE:
When using EPA/incandescent sensors (PN 801080002EPA):
n DO NOT power up the instrument without the sensor connected
n DO NOT disconnect and reconnect a sensor while an analyzer is powered
If this is inconvenient or cannot be avoided:
3. Cycle power to the instrument after connecting the sensor or...
4. Perform a Slope Calibration or Standard Calibration routine after connecting the sensor.
Following these guidelines will extend the life of the incandescent lamp and avoid premature warnings and faults
due to reduced lamp life.
16
MODEL CLARITY II
SECTION 3.0
WIRING
FIGURE 36 Power Wiring for Model 1056 85265 VAC Power Supply (03 ordering code)
To Main PCB
FIGURE 37 Power Wiring for Model 1056 24VDC Power Supply (02 ordering code)
17
MODEL CLARITY II
SECTION 3.0
WIRING
FIGURE 38 Output Wiring for Model 1056 Main PCB
18
MODEL CLARITY II
SECTION 4.0
DISPLAY AND OPERATION
SECTION 4.0
DISPLAY AND OPERATION
4.1
4.2
4.3
4.4
USER INTERFACE
KEYPAD
MAIN DISPLAY
MENU SYSTEM
4.1 USER INTERFACE
The Model 1056 has a large display which shows two
live measurement readouts in large digits and up to four
additional process variables or diagnostic parameters
concurrently. The display is backlit and the format can
be customized to meet user requirements. The intuitive
menu system allows access to Calibration, Hold (of cur
rent outputs), Programming, and Display functions by
pressing the MENU button. In addition, a dedicated
DIAGNOSTIC button is available to provide access to
useful operational information on installed sensor(s)
and any problematic conditions that might occur. The
display flashes Fault and/or Warning when these condi
tions occur. Help screens are displayed for most fault
and warning conditions to guide the user in trou
bleshooting.
During calibration and programming, key presses cause
different displays to appear. The displays are self
explanatory and guide the user stepbystep through
the procedure.
4.2 INSTRUMENT KEYPAD
There are 4 Function keys and 4 Selection keys on the
instrument keypad.
Function keys:
The MENU key is used to access menus for program
ming and calibrating the instrument. Four toplevel
menu items appear when pressing the MENU key:
Calibrate: calibrate attached sensors and
analog outputs.
Hold: Suspend current outputs.
Program: Program outputs, measurement,
temperature, security and reset.
Display: Program display format, language,
warnings, and contrast
Pressing MENU always causes the main menu screen
to appear. Pressing MENU followed by EXIT causes
the main display to appear.
19
MODEL CLARITY II
Pressing the DIAG key displays active Faults and
Warnings, and provides detailed instrument information
and sensor diagnostics including: Faults, Warnings,
Sensor 1 and 2 information, Out 1 and Out 2 live current
values, model configuration string e.g. 1056012031
AN, Instrument Software version, and AC frequency
used. Pressing ENTER on Sensor 1 or Sensor 2 pro
vides useful diagnostics and information (as applica
ble): Measurement, Sensor Type, Raw signal value,
Cell constant, Zero Offset, Temperature, Temperature
SECTION 4.0
DISPLAY AND OPERATION
Offset, selected measurement range, Cable
Resistance, Temperature Sensor Resistance, Signal
Board software version.
The ENTER key. Pressing ENTER stores numbers and
settings and moves the display to the next screen.
The EXIT key. Pressing EXIT returns to the previous
screen without storing changes.
Selection keys:
Surrounding the ENTER key, four Selection keys – up,
down, right and left, move the cursor to all areas of the
screen while using the menus.
Selection keys are used to:
1. select items on the menu screens
2. scroll up and down the menu lists.
3. enter or edit numeric values.
4. move the cursor to the right or left
5. select measurement units during operations
4.3 MAIN DISPLAY
The Model 1056 displays one or two primary measurement
values, up to four secondary measurement values, a
fault and warning banner, alarm relay flags, and a
digital communications icon.
Process measurements:
Two process variables are displayed if two signal boards are installed. One process variable and process tem
perature is displayed if one signal board is installed with one sensor. The Upper display area shows the Sensor
1 process reading. The Center display area shows the Sensor 2 process reading.
For single input configurations, the Upper display area
shows the live process variable.
Secondary values:
Up to four secondary values are shown in four display
quadrants at the bottom half of the screen. All four
secondary value positions can be programmed by the
user to any display parameter available. Possible
secondary values include:
Displayable Secondary Values
Slope 1
Man Temp 2
Ref Off 1
Output 1 mA
Gl Imp 1
Output 2 mA
Ref Imp 1
Output 1 %
Raw
Output 2 %
mV Input
Measure 1
Temp 1
Blank
Man Temp 1
20
MODEL CLARITY II
SECTION 4.0
DISPLAY AND OPERATION
Fault and Warning banner:
If the analyzer detects a problem with itself or the sensor the word Fault or Warning will appear at the bottom of
the display. A fault requires immediate attention. A warning indicates a problematic condition or an impending fail
ure. For troubleshooting assitance, press Diag.
Formatting the Main Display
The main display screen can be programmed to show primary process variables, secondary process variables and
diagnostics.
1.
Press MENU
2.
Scroll down to Display. Press ENTER.
3.
Main Format will be highlighted. Press ENTER.
4.
The sensor 1 process value will be highlighted in reverse video. Press the selection keys to navigate down
to the screen sections that you wish to program. Press ENTER.
5.
Choose the desired display parameter or diagnostic for each of the four display sections in the lower screen.
6.
Continue to navigate and program all desired screen sections.
return to the main display.
Press MENU and EXIT. The screen will
For single sensor configurations, the default display shows the live process measurement in the upper display area
and temperature in the center display area. The user can elect to disable the display of temperature in the cen
ter display area using the Main Format function. See Fig. 41 to guide you through programming the main display
to select process parameters and diagnostics of your choice.
For dual sensor configurations, the default display shows Sensor 1 live process measurement in the upper display
area and Sensor 2 live process measurement temperature in the center display area. See Fig. 41 to guide you
through programming the main display to select process parameters and diagnostics of your choice.
4.4
MENU SYSTEM
Model 1056 uses a scroll and select menu system.
Pressing the MENU key at any time opens the toplevel
menu including Calibrate, Hold, Program and Display
functions.
To find a menu item, scroll with the up and down keys
until the item is highlighted. Continue to scroll and
select menu items until the desired function is chosen.
To select the item, press ENTER. To return to a previ
ous menu level or to enable the main live display,
press the EXIT key repeatedly. To return immediately
to the main display from any menu level, simply press
MENU then EXIT.
The selection keys have the following functions:
The Up key (above ENTER) increments numerical values, moves the decimal place one place to the right,
or selects units of measurement.
The Down key (below ENTER) decrements numerical values, moves the decimal place one place to the
left, or selects units of measurement
The Left key (left of ENTER) moves the cursor to the left.
The Right key (right of ENTER) moves the cursor to the right.
To access desired menu functions, use the “Quick Reference” Figure B. During all menu displays (except main
display format and Quick Start), the live process measurements and secondary measurement values are
displayed in the top two lines of the Upper display area. This conveniently allows display of the live values during
important calibration and programming operations.
Menu screens will time out after two minutes and return to the main live display.
21
MODEL CLARITY II
SECTION 4.0
DISPLAY AND OPERATION
FIGURE 41 Formatting the Main Display
22
MODEL CLARITY II
SECTION 5.0
PROGRAMMING THE ANALYZER BASICS
SECTION 5.0.
PROGRAMMING THE ANALYZER BASICS
5.1
5.2
5.3
5.4
5.5
5.6
5.7
5.8
GENERAL
CHANGING STARTUP SETTINGS
CONFIGURING AND RANGING 420MA OUTPUTS
SETTING SECURITY CODES
SECURITY ACCESS
USING HOLD
RESETTING FACTORY DEFAULTS – RESET ANALYZER
PROGRAMMING ALARM RELAYS
5.1 GENERAL
Section 5.0 describes the following programming functions:
Changing the measurement type, measurement units and temperature units.
Configure and assign values to the current outputs
Set a security code for two levels of security access
Accessing menu functions using a security code
Enabling and disabling Hold mode for current outputs
Resetting all factory defaults, calibration data only, or current output settings only
5.2 CHANGING STARTUP SETTINGS
5.2.1 Purpose
To change the measurement type, measurement units, or temperature units that were initially entered in Quick
Start, choose the Reset analyzer function (Sec. 5.7) or access the Program menus for sensor 1 or sensor 2 (Sec.
6.0). The following choices for specific measurement type, measurement units are available for each sensor meas
urement board.
5.2.2 Procedure.
Follow the Reset Analyzer procedure (Sec 5.7) to reconfigure the analyzer to display new measurements or
measurement units. To change the specific measurement or measurement units for each signal board type,
refer to the Program menu for the appropriate measurement (Sec. 6.0).
5.3 CONFIGURING AND RANGING THE CURRENT OUTPUTS
5.3.1 Purpose
The Model 1056 accepts inputs from two sensors and
has two analog current outputs. Ranging the outputs
means assigning values to the low (0 or 4 mA) and high
(20 mA) outputs. This section provides a guide for
configuring and ranging the outputs. ALWAYS
CONFIGURE THE OUTPUTS FIRST.
5.3.2 Definitions
1. CURRENT OUTPUTS. The analyzer provides a con
tinuous output current (420 mA or 020 mA) directly
proportional to the process variable or temperature.
The low and high current outputs can be set to any
value.
2. ASSIGNING OUTPUTS. Assign a measurement to
Output 1 or Output 2.
3. DAMPEN. Output dampening smooths out noisy
readings. It also increases the response time of the
output. Output dampening does not affect the
response time of the display.
4. MODE. The current output can be made directly
proportional to the displayed value (linear mode) or
directly proportional to the common logarithm of the
displayed value (log mode).
23
MODEL CLARITY II
SECTION 5.0
PROGRAMMING THE ANALYZER BASICS
5.3.3. Procedure: Configure Outputs.
Under the Program/Outputs menu, the adjacent screen
will appear to allow configuration of the outputs. Follow
the menu screens in Fig. 52 to configure the outputs.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
OutputM Configure
Assign:
S1 Meas
Range:
420mA
Scale:
Linear
Dampening:
0sec
Fault Mode: Fixed
Fault Value: 21.00mA
5.3.4. Procedure: Assigning Measurements the Low
and High Current Outputs
The adjacent screen will appear when entering the
Assign function under Program/Output/Configure.
These screens allow you to assign a measurement,
process value, or temperature input to each output.
Follow the menu screens in Fig. 52 to assign
measurements to the outputs.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
5.3.5. Procedure: Ranging the Current Outputs
The
adjacent
screen
will
appear
under
Program/Output/Range. Enter a value for 4mA and
20mA (or 0mA and 20mA) for each output. Follow the
menu screens in Fig. 52 to assign values to the out
puts.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
OutputM Assign
S1 Measurement
S2 Measurement
OM
OM
OM
OM
SN
SN
SN
SN
Output Range
4mA: 0.000µS/cm
20mA: 20.00µS/cm
4mA: 00.00pH
20mA: 14.00pH
Figure 52. Configuring and Ranging the Current Outputs
24
MODEL CLARITY II
SECTION 5.0
PROGRAMMING THE ANALYZER BASICS
2.
3.
5.4 SETTING A SECURITY CODE
5.4.1 Purpose.
The security codes prevent accidental or unwanted
changes to program settings, displays, and calibration.
Model 1056 has two levels of security code to control
access and use of the instrument to different types of
users. The two levels of security are:
All: This is the Supervisory security level. It
allows access to all menu functions, including
Programming, Calibration, Hold and Display.
Calibration/Hold: This is the operator or
technician level menu. It allows access to
only calibration and Hold of the current outputs.
4.
Scroll down to Security. Select Security.
The security entry screen appears. Enter a
three digit security code for each of the desired
security levels. The security code takes effect
two minutes after the last key stroke. Record
the security code(s) for future access and
communication to operators or technicians as
needed.
The display returns to the security menu
screen. Press EXIT to return to the previous
screen. To return to the main display, press
MENU followed by EXIT.
Fig. 53 displays the security code screens.
5.4.2 Procedure.
1. Press MENU. The main menu screen appears.
Choose Program.
Program
MAIN MENU
Figure 53. Setting a Security Code
S1: 1.234µS/cm
S2: 12.34pH
25.0ºC
25.0ºC
Program
Outputs
Measurement
Temperature
S1: 1.234µS/cm
S2: 12.34pH
Security
Diagnostic Setup
Ambient AC Power:Unk
Reset Analyzer
25.0ºC
25.0ºC
Security
Calibration/Hold: 000
All: 000
25
MODEL CLARITY II
SECTION 5.0
PROGRAMMING THE ANALYZER BASICS
5.5 SECURITY ACCESS
5.6 USING HOLD
5.5.1 How the Security Code Works
When entering the correct access code for the
Calibration/Hold security level, the Calibration and
Hold menus are accessible. This allows operators or
technicians to perform routine maintenance. This
security level does not allow access to the Program or
Display menus.
When entering the correct access code for All security
level, the user has access to all menu functions, includ
ing Programming, Calibration, Hold and Display.
5.6.1 Purpose
The analyzer output is always proportional to measured
value. To prevent improper operation of systems or
pumps that are controlled directly by the current
output, place the analyzer in hold before removing
the sensor for calibration and maintenance. Be sure
to remove the analyzer from hold once calibration is
complete. During hold, both outputs remain at the last
value. Once in hold, all current outputs remain on
Hold indefinitely.
5.5.2 Procedure.
1. If a security code has been programmed, selecting
the Calibrate, Hold, Program or Display top menu
items causes the security access screen to appear
5.6.2 Using the Hold Function
To hold the outputs,
1.
Press MENU. The main menu screen appears.
Choose Hold.
2.
The Hold Outputs and Alarms? screen
appears. Choose Yes to place the analyzer in
hold. Choose No to take the analyzer out of
hold.
Note: There are no alarm relays with this con
figuration. Current outputs are included with all
configurations.
3.
The Hold screen will then appear and Hold
will remain on indefinitely until Hold is
disabled.
2. Enter the threedigit security code for the appropriate
security level.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
Security Code
000
3. If the entry is correct, the appropriate menu screen
appears. If the entry is incorrect, the Invalid Code
screen appears. The Enter Security Code screen
reappears after 2 seconds.
See figure 51 below.
26
Hold
MAIN MENU
Figure 54. Using Hold
S1: 1.234µS/cm
S2: 12.34pH
25.0ºC
25.0ºC
Hold
S1 Hold:
S2 Hold:
No
No
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
S1 Hold outputs
and alarms?
No
Yes
MODEL CLARITY II
SECTION 5.0
PROGRAMMING THE ANALYZER BASICS
5.7 RESETTING FACTORY DEFAULT SETTINGS
5.7.1 Purpose.
This section describes how to restore factory calibration and default values. The process also clears all fault messages
and returns the display to the first Quick Start screen. The Model 1056 offers three options for resetting factory
defaults.
a. reset all settings to factory defaults
b. reset sensor calibration data only
c. reset analog output settings only
5.7.2. Procedure.
To reset to factory defaults, reset calibration data only or reset analog outputs only, follow the Reset Analyzer flow
diagram.
Figure 55. Resetting Factory Default Settings
27
MODEL CLARITY II
SECTION 5.0
PROGRAMMING THE ANALYZER BASICS
5.8 Programming Alarm Relays
5.8.1 Purpose.
The Model 1056 24VDC (02 order code) and the AC switching power supply (03 order code) provide four alarm
relays for process measurement or temperature. Each alarm can be configured as a fault alarm instead of a
process alarm. Also, each relay can be programmed independently and each can be programmed as an interval
timer. This section describes how to configure alarm relays, simulate relay activation, and synchronize timers for
the four alarm relays. This section provides details to program the following alarm features:
Sec.
Alarm relay feature:
default
Description
5.9.2
Enter Setpoint
100.0uS/cm
Enter alarm trigger value
5.9.3
Assign measurement S1 Measure
Select alarm assignment
5.9.4
Set relay logic
Program relay to activate at High or Low reading
5.9.5
Deadband:
High
0.00uS/cm
5.9.6
Normal state:
5.9.7
Interval time:
24.0 hr
Time in hours between relay activations
5.9.8
OnTime:
10 min
Enter the time in seconds that the relay is activated.
5.9.9
Recover time:
60 sec
Enter time after the relay deactivation for process recovery
5.9.10
Hold while active:
5.9.11
Simulate
5.9.12
Synchronize Timers
Open
Program the change in process value after the relay deactivates
S1
Program relay default condition as open or closed for failsafe operation
Holds current outputs during relay activation
Manually simulate alarms to confirm relay operation
Yes
Control the timing of two or more relay timers set as Interval timers
Under the Program/Alarms menu, this screen will
appear to allow configuration of the alarm relays.
Follow the menu screens in Fig. XX to configure the
outputs.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
This screen will appear to allow selection of a specific
alarm relay. Select the desired alarm and press
ENTER.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
This screen will appear next to allow complete pro
gramming of each alarm. Factory defaults are dis
played as they would appear for an installed contact
ing conductivity board. USP Safety only appears if
alarm logic is set to “USP”. Interval timer, On Time,
Recover Time, and Hold While Active only appear if
the alarm is configured as an Interval timer.
Alarms
Configure/Setpoint
Simulate
Synchronize Timers: Yes
Configure/Setpoint
Alarm 1
Alarm 2
Alarm 3
Alarm 4
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
AlarmM Settings
Setpoint:
100.0uS/cm
Assign:
S1 Measure
Logic:
High
Deadband: 0.00uS/cm
USP Safety:
0%↓
Interval time:
24.0 hr
On Time:
120 sec
Recover time:
60 sec
Hold while active: Sens1
28
MODEL CLARITY II
5.8.2 Procedure – Enter Setpoints
Under the Program/Alarms menu, this screen will
appear to allow configuration of the alarm relays.
Enter the desired value for the process measurement
or temperature at which to activate an alarm event.
5.8.3 Procedure – Assign Measurement
Under the Alarms Settings menu, this screen will
appear to allow assignment of the alarm relays. select
an alarm assignment. Additional assignment choices
are shown in Figure XX depending on which meas
urement board(s) is installed.
SECTION 5.0
PROGRAMMING THE ANALYZER BASICS
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
Alarm1 S2 Setpoint
+100.0uS/cm
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
AlarmM Assign:
S1 Measurement
S2 Measurement
Interval Timer
Fault
Off
5.8.4 Procedure – Set Relay Logic
Under the Alarms Settings menu, this screen will
appear to set the alarm logic. Select the desired relay
logic to activate alarms at a High reading or a Low
reading. USP Safety only appears if a contacting con
ductivity board is installed.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
5.8.5 Procedure – Deadband
Under the Alarms Settings menu, this screen will
appear to program the deadband as a measurement
value. Enter the change in the process value needed
after the relay deactivates to return to normal (and
thereby preventing repeated alarm activation).
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
5.8.6 Procedure – Normal state
The user can define failsafe condition in software by
programming the alarm default state to normally open
or normally closed upon power up. To display this
alarm configuration item, enter the Expert menus by
holding down the EXIT key for 6 seconds while in the
main display mode. Select Yes upon seeing the screen
prompt: “Enable Expert Menu?”
Under the Alarms Settings menu, this screen will
appear to set the normal state of the alarms. Select the
alarm condition that is desired each time the analyzer is
powering up.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
5.8.7 Procedure – Interval time
Under the Alarms Settings menu, this screen will
appear to set the interval time. Enter the fixed time in
hours between relay activations.
AlarmM Logic:
High
Low
USP
Alarm1 Deadband
+000.5uS/cm
Alarm2 Normal State
Open
Closed
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
Alarm1 Interval Time
024.0 hrs
29
MODEL CLARITY II
5.8.8 Procedure – On time
Under the Alarms Settings menu, this screen will
appear to set the relay on time. Enter the time in sec
onds that the relay is activated.
5.8.9 Procedure – Recovery time
Under the Alarms Settings menu, this screen will
appear to set the relay recovery time. Enter time after
the relay deactivation for process recovery.
SECTION 5.0
PROGRAMMING THE ANALYZER BASICS
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
Alarm1 OnTime
00.00sec
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
Alarm1 Recovery
060sec
5.8.10 Procedure – Hold while active
Under the Alarms Settings menu, this screen will
appear to program the feature that Holds the current
outputs while alarms are active. Select to hold the
current outputs for Sensor 1, Sensor 2 or both sensors
while the relay is activated.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
5.8.11 Procedure – Simulate
Alarm relays can be manually set for the purposes of
checking devices such as valves or pumps. Under the
Alarms Settings menu, this screen will appear to allow
manual forced activation of the alarm relays. Select
the desired alarm condition to simulate.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
5.8.12 Procedure – Synchronize
Under the Alarms Settings menu, this screen will
appear to allow Synchronization of alarms that are set
to interval timers. Select yes or no to Synchronize
two or more timers.
30
Alarm1 Hold while active
Sensor 1
Sensor 2
Both
None
Simulate Alarm M
Don’t simulate
Deenergize
Energize
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
Synchronize Timers
Yes
No
MODEL CLARITY II
SECTION 6.0
TURBIDITY
SECTION 6.0
PROGRAMMING TURBIDITY
6.1
6.2
6.3
6.4
CONFIGURING MEASUREMENTS – INTRODUCTION
TURBIDITY
CHOOSING TURBIDITY OR TOTAL SUSPENDED SOLIDS
ENTERING A TURBIDITY TO TSS CONVERSION EQUATION
6.1 PROGRAMMING MEASUREMENTS – INTRODUCTION
The Model 1056 automatically recognizes each installed measurement board upon first powerup and each time
the analyzer is powered. Completion of Quick Start screens upon first power up enable measurements, but addi
tional steps may be required to program the analyzer for the desired measurement application. This section cov
ers the following programming and configuration functions;
1. Selecting measurement type or sensor type (all sections)
2. Defining measurement display units (all sections)
3. Adjusting the input filter to control display and output reading variability or noise (all sections)
4. Entering TSS data
5. Information on bubble rejection alogorithm
To fully configure the analyzer for each installed measurement board, you may use the following:
1. Reset Analyzer function to reset factory defaults and configure the measurement board to the desired
measurement. Follow the Reset Analyzer menu to reconfigure the analyzer to display new measurements
or measurement units.
2. Program menus to adjust any of the programmable configuration items. Use the following configuration
and programming guidelines for the applicable measurement.
31
MODEL CLARITY II
SECTION 6.0
TURBIDITY
6.2 TURBIDITY MEASUREMENT PROGRAMMING
6.2.1 DESCRIPTION
This section describes how to configure the Model 1056 analyzer for Turbidity measurements. The following
programming and configuration functions are covered.
TABLE 611 TURBIDITY MEASUREMENT PROGRAMMING
Measure
Turbidity
Sec.
Menu function:
6.9.2
Measurement type:
Turbidity
6.9.3
Measurement units:
NTU
6.9.4
Enter TSS* Data:
6.9.5
Filter:
6.9.6
Bubble Rejection:
default
Description
Select Turbidity or TSS calculation (estimated TSS)
NTU, FTU, FNU
Enter TSS and NTU data to calculate TSS based on Turbidity
20 sec
On
Override the default input filter, enter 0999 seconds
Intelligent software algorithm to eliminate erroneous readings
caused by bubble accumulation in the sample
*TSS: Total Suspended Solids
A detailed flow diagram for Turbidity programming is provided at the end of Sec. 6 to guide you through all basic
programming and configuration functions.
To
1.
2.
3.
4.
configure the Turbidity measurement board:
Press MENU
Scroll down to Program. Press ENTER.
Scroll down to Measurement. Press ENTER.
Select Sensor 1 or Sensor 2 corresponding to Turbidity. Press ENTER.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Configure
Measure:
Turbidity
Units:
NTU
Enter TSS Data
Filter:
20sec
Bubble Rejection: On
The following screen format will appear (factory defaults are shown).
To program Turbidity, scroll to the desired item and press ENTER.
The following subsections provide you with the initial display screen that appears for each programming routine.
Use the flow diagram for Turbidity programming at the end of Sec. 6 and the live screen prompts to complete
programming.
6.2.2 Measurement
The display screen for selecting the measurement is
shown. The default measurement is displayed in bold
type. Refer to the Turbidity Programming flow diagram
to complete this function.
6.2.3 Units
The display screen for selecting the measurement
units is shown. The default value is displayed in bold
type. Refer to the Turbidity Programming flow diagram
to complete this function.
32
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Measurement
Turbidity
Calculated TSS
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Units
NTU
FTU
FNU
MODEL CLARITY II
SECTION 6.0
TURBIDITY
If TSS data (Total Suspended Solids) calculation is
selected, the following screen will be displayed. Refer
to the Turbidity programming flow diagram to complete
this function.
6.2.4 Enter TSS Data
The display screen for entering TSS Data is shown.
The default values are displayed. Refer to the
Turbidity Programming flow diagram to complete this
function
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Units
ppm
mg/L
none
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN TSS Data
Pt1 TSS:
0.000ppm
Pt1 Turbid: 0.000NTU
Pt2 TSS:
100.0ppm
Pt2 Turbid: 100.0NTU
Calculate
Note: Based on userentered NTU data, calculating
TSS as a straight line curve could cause TSS to go
below zero. The following screen lets users know that
TSS will become zero below a certain NTU value.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN TSS Data
Calculation Complete
Calculated TSS = 0 below
xxxx NTU
The following illustration shows the potential for calculated TSS to go below zero
Normal case: TSS is always a positive number when Turbidity is a positive number.
Abnormal case: TSS can be a negative number when Turbidity is a positive number.
33
MODEL CLARITY II
When the TSS data entry is complete, press ENTER.
The display will confirm the determination of a TSS
straight line curve fit to the entered NTU/turbidity data
by displaying this screen:
The following screen may appear if TSS calculation is
unsuccessful. Reentry of NTU and TSS data is
required.
SECTION 6.0
TURBIDITY
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN TSS Data
Calculation
Complete
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN TSS Data
Data Entry Error
Press EXIT
6.2.5 Filter
The display screen for entering the input filter value in
seconds is shown. The default value is displayed in
bold type. Refer to the Turbidity Programming flow
diagram to complete this function.
6.2.6 Bubble Rejection
Bubble rejection is an internal software algorithm that
characterizes turbidity readings as bubbles as
opposed to true turbidity of the sample. With Bubble
rejection enabled, these erroneous readings are elimi
nated from the live measurements shown on the dis
play and transmitted via the current outputs.
The display screen for selecting bubble rejection algo
rithm is shown. The default setting is displayed in bold
type. Refer to the Turbidity Programming flow diagram
to complete this function.
34
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Input Filter
020sec
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Bubble Rejection
On
Off
MODEL CLARITY II
SECTION 6.0
TURBIDITY
6.3 CHOOSING TURBIDITY OR TOTAL SUSPENDED SOLIDS
6.3.1 Purpose
This section describes how to do the following:
1. Configure the analyzer to display results as turbidity or total suspended solids (TSS).
2. Choose units in which results are to be displayed.
3. Select a time period for signal averaging.
4. Enable or disable bubble rejection software.
6.3.2 Definitions
1. TURBIDITY. Turbidity is a measure of the
amount of light scattered by particles in a sam
ple. Figure 64 illustrates how turbidity is meas
ured. A beam of light passes through a sample
containing suspended particles. The particles
interact with the light and scatter it in all direc
tions. Although the drawing implies scattering
is equal in all directions, this is generally not the
case. For particles bigger than about 1/10 of
the wavelength of light, scattering is highly
directional. A detector measures the intensity of
scattered light.
FIGURE 64. Turbidity Sensor — General
Measured turbidity is dependent on instrumen
tal conditions. In an attempt to allow turbidities
measured by different instruments to be com
pared, two standards for turbidity instruments
have evolved. USEPA established Method
180.1, and the International Standards
Organization established ISO 7027. EPA
Method 180.1 must be used for reporting pur
poses in the United States. Figure 65 shows
an EPA 180.1 turbidimeter. Figure 66 shows
an ISO 7027 turbidimeter.
EPA Method 180.1 requires that:
FIGURE 65. Turbidity Sensor — EPA 180.1
A. The light source be a tungsten lamp operat
ed with a filament temperature between 2200 and 2700 K.
B. The detector have optimum response between 400 and 600 nm (approximates the human eye).
C. The scattered light be measured at 90º±30º with respect to the incident beam.
D. The total path length of the light through the sample be less than 10 cm.
Requirements A and B essentially restrict the measurement to visible light. Although the most of the energy
radiated by an incandescent lamp is in the near infrared, keeping the filament temperature between 2200 and
2700 K, ensures that at least some energy is available in the visible range. Further specifying that the detec
tor and filter combination have maximum sensitivity between 400 nm (violet light) and 600 nm (orange light),
cements the measurement in the visible range. Wavelength is important because particles scatter light most
efficiently if their size is approximately equal to the wavelength of light used for the measurement. The longer
the wavelength, the more sensitive the measurement is to larger diameter particles and the less sensitive it is
to smaller diameter particles.
continued on following page
35
MODEL CLARITY II
SECTION 6.0
TURBIDITY
Requirement C is arbitrary. The light scattered by a particle depends on the shape and size of the particle, the
wavelength used for the measurement, and the angle of observation. Choosing 90º avoids the difficulties of
having to integrate the scattered light over all the scattering angles. An arbitrary observation angle works so
long as the sample turbidity is referred to the turbidity of a standard solution measured at the same angle. A
turbidimeter that measures scattered light at 90º is called a nephelometer.
Requirement D has a lot to do with the linearity of the
sensor. As Figures 65 and 66 show, particles lying
between the measurement zone and the detector can
scatter the scattered radiation. This secondary scatter
ing reduces the amount of light striking the detector. The
result is a decrease in the expected turbidity value and
a decrease in linearity. The greater the amount of sec
ondary scattering, the greater the nonlinearity. Particles
in the area between the source and measurement zone
also reduce linearity.
ISO 7027 requirements are somewhat different from
EPA requirements. ISO 7027 requires that:
FIGURE 66. Turbidity Sensor — ISO 7027
A. The wavelength of the interrogating light be 860±60nm, or for colorless samples, 550±30nm.
B. The measuring angle be 90±2.5º.
ISO 7027 does not restrict the maximum light path length through the sample. ISO 7027 calls out beam geom
etry and aperture requirements that EPA 180.1 does not address.
Although ISO 7027 allows a laser, light emitting diode, or tunsten filament lamp fitted with an interference fil
ter as the light source, most instruments, including the Clarity II, use an 860 nm LED. Because ISO 7027 tur
bidimeters use a longer wavelength for the measurement, they tend to be more sensitive to larger particles
than EPA 180.1 turbidimeters. Turbidities measured using the EPA and ISO methods will be different.
2. TOTAL SUSPENDED SOLIDS. Total suspended solids (TSS) is a measure of the total mass of particles in a
sample. It is determined by filtering a volume of sample and weighing the mass of dried residue retained on
the filter. Because turbidity arises from suspended particles in water, turbidity can be used as an alternative
way of measuring total suspended solids (TSS). The relation between turbidity and TSS is wholly empirical and
must be determined by the user.
3. TURBIDITY UNITS. Turbidity is measured in units of NTU (nephelometric turbidity units), FTU (formazin tur
bidity units), or FNU (formazin nephelometric units). Nephelometry means the scattered light is measured at
90º to the interrogating beam. Formazin refers to the polymer suspension typically used to calibrate turbidity
sensors. The units — NTU, FTU, and FNU — are equivalent.
4. TSS UNITS. The TSS value calculated from the turbidity measurement can be displayed in units of ppm or
mg/L. The user can also choose to have no units displayed.
5. SIGNAL AVERAGING. Signal averaging is a way of filtering noisy signals. Signal averaging reduces random
fluctuation in the signal but increases the response time to step changes. Recommended signal averaging is
20 sec. The reading will take 20 seconds to reach 63% of its final value following a step change greater than
the filter threshold.
6. BUBBLE REJECTION. When a bubble passes through the light beam, it reflects light onto the measuring pho
todiode, causing a spike in the measured turbidity. The Model 1056 analyzer has proprietary software that
rejects the turbidity spikes caused by bubbles.
36
MODEL CLARITY II
SECTION 6.0
TURBIDITY
6.3.3 Procedure: Selecting Turbidity or TSS
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
1. Press MENU. The main menu screen appears. Choose Program.
2. Choose Measurement.
3. Choose Sensor 1 or Sensor 2. For a single input configuration, the Sensor 1 Sensor 2 screen does not
appear.
4. Choose Turbidity or TSS.
5. Choose the desired units:
a. For turbidity choose NTU, FTU, or FNU.
b. For TSS choose ppm, mg/L, or none.
6. Choose Bubble Rejection.
7. Choose On to enable bubble rejection software. Choose Off to disable.
8. Press EXIT to return to the previous screen. To return to the main display, press MENU followed by EXIT.
6.4 ENTERING A TURBIDITY TO TSS CONVERSION EQUATION
6.4.1 Purpose
The analyzer can be programmed to convert turbidity to a total suspended solids (TSS) reading. There is no fun
damental relationship between turbidity and TSS. Every process stream is unique. The user must determine the
relationship between turbidity and TSS for his process. The analyzer accepts only a linear calibration curve.
Figure 62 shows how the turbidity to TSS conversion works. The user enters two points P1 and P2, and the ana
lyzer calculates the equation of a straight line between the points. The analyzer then converts all subsequent tur
bidity measurements to TSS using the equation. It is important to note that if the cause or the source of the tur
bidity changes, new points P1 and P2 will need to be determined and the calibration repeated.
The accuracy of the measurement depends on how linear the actual relationship between TSS and turbidity is. At
a minimum, the user should confirm linearity by diluting the most turbid sample (P2) and verifying that the new tur
bidity and TSS point lies reasonably close to the line. Ideally, the dilution should be done with filtered sample, not
deionized water. Deionized water can change the index of refraction of the liquid and can increase or decrease the
solubility of the particles. Therefore, the diluted sample will not be representative of the process liquid. For a more
rigorous procedure for checking linearity and developing values to enter for points P1 and P2, refer to the
Appendix.
After the analyzer has calculated the turbidity to TSS conversion equation, it also calculates the xintercept (NTU).
See Figure 63. If the xintercept is greater than zero, the analyzer will display that value as the lowest turbidity
reading it will accept. A lower turbidity reading will produce a negative TSS value. If the xintercept in less than
zero, the screen does not appear.
37
MODEL CLARITY II
SECTION 6.0
TURBIDITY
6.4 ENTERING A TURBIDITY TO TSS CONVERSION EQUATION cont.
FIGURE 62. Converting Turbidity to TSS
FIGURE 63. Lowest Turbidity (TSS)
38
MODEL CLARITY II
SECTION 6.0
TURBIDITY
6.4.2 Procedure
1. First, calibrate the sensor. See Section Section 6.2, 6.3, or 6.4.
2. Press the MENU key. The main menu appears. Choose Program.
3. Choose Measurement
4. Choose Sen1 (sensor 1) or Sen2 (sensor 2).
5. Choose Enter TSS Data.
6. The display prompts the user to enter TSS for point 1 (Pt1). The units shown in the second line will be the units
selected in Section 5.5.3. Press ENTER.
7. The display prompts the user to enter the turbidity for point 1. Press ENTER.
8. The display prompts the user to enter TSS for point 2 (Pt2). Press ENTER.
9. The display prompts the user to enter the turbidity for point 2. Press ENTER.
10. The screen at left appears if the calibration was successful. Press ENTER.
11. If the calibration was unsuccessful, repeat steps 6 through 9, checking for data entry errors.
12. If the intercept on the NTU axis is negative, the analyzer will display the low turbidity limit.
13. To return to the main display, press MENU then EXIT.
39
FIGURE 67 Configure Turbidity Measurement
MODEL CLARITY II
40
SECTION 6.0
TURBIDITY
MODEL CLARITY II
SECTION 7.0
CALIBRATION
SECTION 7.0
CALIBRATION
7.1 CALIBRATION – INTRODUCTION
7.2 TURBIDITY CALIBRATION
7.1 CALIBRATION – INTRODUCTION
Calibration is the process of adjusting or standardizing
the analyzer to a lab test or a calibrated laboratory
instrument, or standardizing to some known reference
(such as a commercial buffer).
The autorecognition feature of the analyzer will enable
the appropriate calibration screens to allow calibration for
any single sensor configuration or dual sensor config
uration of the analyzer. Completion of Quick Start upon
first power up enables live measurements but does not
ensure accurate readings in the lab or in process.
Calibration should be performed with each attached
sensor to ensure accurate, repeatable readings.
7.2 TURBIDITY CALIBRATION
7.2.1 DESCRIPTION
This section describes how to calibrate the turbidity sensor against a userprepared standard as a 2point calibra
tion with diionized water, against a 20 NTU userprepared standard as a single point calibration, and against a
grab sample using a reference turbidimeter.
THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1056 WITH AN ATTACHED TURBIDITY
SENSOR AS PART OF THE COMPLETE CLARITY II TURBIDITY SYSTEM. THE FOLLOWING CALIBRATION
ROUTINES ARE COVERED.
TABLE 712 TURBIDITY CALIBRATION ROUTINES
Measure Sec. Calibration function:
Turbidity 7.10.2 Slope Calibration
default value
Description
Slope cal with pure water and a standard of known turbidity
7.10.3 Standardize Calibration
Standardizing the sensor to a known turbidity
7.10.4 Grab Calibration
Standardizing the sensor to a known turbidity based on a
reference turbidimeter
A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines.
To calibrate Turbidity:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to
Turbidity. Press ENTER.
4. Select Turbidity. Press ENTER.
The following subsections provide you with the initial
display screen that appears for each calibration rou
tine. Use the flow diagram for Turbidity calibration
at the end of Sec. 7 and the live screen prompts to
complete calibration.
The following screen will appear.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibrate?
Turbidity
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Calibrate
Slope
Standard
Grab
41
MODEL CLARITY II
7.2.2 SLOPE CALIBRATION — Turbidity
This section describes how to conduct a 2point cali
bration of the turbidity sensor against a userprepared
20NTU standard. The calibration requires two steps.
First, immerse the sensor in filtered water having very
low turbidity and measure the sensor output. Next,
increase the turbidity of the filtered water by a known
amount, typically 20 NTU, and measure the sensor
output again. The analyzer takes the two measure
ments, applies a linearization correction (if necessary),
and calculates the sensitivity. Sensitivity is the sensor
output (in mV) divided by turbidity. A typical new sen
sor has a sensitivity of about 10 mV/NTU. As the sen
sor ages, the sensitivity decreases. The figure below
illustrates how turbidity calibration works. Before
beginning the calibration, the analyzer does a dark
current measurement. Dark current is the signal gener
ated by the detector when no light is falling on it. The
analyzer subtracts the dark current from the raw scat
This screen appears after selecting Slope calibration.
SECTION 7.0
CALIBRATION
tered light signal and converts the result to turbidity. In
highly filtered samples, which scatter little light, the
dark current can be a substantial amount of the signal
generated by the detector.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Slope Cal
Sensor in pure H2O?
Press ENTER
The following screen will appear if Slope Cal is suc
cessful. The screen will return to the Turbidity Cal
Menu.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
The following screen may appear if Slope Cal is
unsuccessful.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Slope Cal
Cal Complete
SN Slope Cal
Calibration
Error
Press EXIT
42
MODEL CLARITY II
7.2.3 STANDARDIZE CALIBRATION Turbidity
The turbidity sensor can also be calibrated against a
commercial standard. Stable 20.0 NTU standards are
available from a number of sources. Calibration using a
commercial standard is simple. Filtered deionized water
is not required. Before beginning the calibration, the
analyzer does a dark current measurement. Dark cur
rent is the signal generated by the detector even when
no light is falling on it. The analyzer subtracts the dark
current from the raw scattered light signal and converts
the result to turbidity. In highly filtered samples, which
scatter little light, the dark current can be a substantial
amount of the signal generated by the sensor.
SECTION 7.0
CALIBRATION
This screen appears after selecting Standard calibration.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Standard Cal
Sensor in Standard?
Press ENTER
The following screen will appear if Standard Cal is
successful. The screen will return to the Turbidity Cal
Menu.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
The following screen may appear if Standard Cal is
unsuccessful.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Standard Cal
Cal Complete
SN Standard Cal
Calibration
Error
Press EXIT
7.2.4 GRAB CALIBRATION Turbidity
If desired, the turbidity sensor can be calibrated
against the turbidity reading from another instrument.
The analyzer treats the value entered by the user as
though it were the true turbidity of the sample.
Therefore, grab sample calibration changes the sensi
tivity, it does not apply an offset to the reading.
The following screen will appear if Grab Cal is successful.
The screen will return to the Turbidity Cal Menu.
This screen appears after selecting Grab calibration.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Grab Cal
Wait for stable
reading
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Grab Cal
Cal Complete
The following screen may appear if Grab Cal is
unsuccessful.
S1: 1.234µS/cm
25.0ºC
S2: 12.34pH
25.0ºC
SN Grab Cal
Calibration
Error
Press EXIT
43
FIGURE 78 Calibrate Turbidity
MODEL CLARITY II
44
SECTION 7.0
CALIBRATION
MODEL CLARITY II
SECTION 8.0
MAINTENANCE
SECTION 8.0
MAINTENANCE
8.1 MODEL 1056
The Model 1056 analyzer used in the Clarity II turbidimeter needs little routine maintenance.
Clean the analyzer case and front panel by wiping it with a clean soft cloth dampened with water ONLY. Do not
use solvents, like alcohol, that might cause a buildup of static charge.
A few of the components of the analyzer are replaceable. See Tables 81 and 82.
WARNING
Explosion Hazard. Do not disconnect equipment when a flammable
or combustible atmosphere is present.
TABLE 81. Replacement Parts for Model 1056
PN
Description
Shipping Weight
2382300
Panel mounting kit, includes four brackets and four set screws
2 lb/1.0 kg
3405900
Gasket, for panel mounting
1 lb/0.5 kg
3406200
Gasket, internal for enclosure
1 lb/0.5 kg
2423000
Hole plug and gland fittings.
2 lb/1.0 kg
Shipping weights are rounded up to the nearest whole lb or 0.5 kg.
45
MODEL CLARITY II
SECTION 8.0
MAINTENANCE
8.2 SENSOR
8.2.1 Cleaning the sensor
Clean the sensor by rinsing it with water followed by wiping with a soft tissue. If water is inadequate, wash with a
mild detergent solution followed by thorough rinsing with water. Do not scratch the lamp or photodiode win
dows.
If mineral scale is present, use a dilute acid solution applied with a cotton swab to clean away the deposit. Rinse
thoroughly with water.
Do not use abrasive cleaners or solvents.
8.2.2 Replacing the lamp/LED board
The USEPAcompliant sensor uses a tungsten filament lamp (PN 109010004EPA) as the light source. The
lamp has an expected life of about one year. The ISOcompliant version uses an infrared LED (PN 10901
0005ISO). Its expected life is five years. The Model 1056 analyzer continuously monitors the source inten
sity and corrects for changes in source intensity caused by age. When the source intensity becomes too low,
the analyzer warns the user. The user should replace the lamp as soon as possible.
To replace the lamp/LED board…
step 4
1. Turn off power to the analyzer.
WARNING!
Explosion hazard. Do not disconnect equipment when a
flammable or combustible atmosphere is present.
2. Remove the sensor from the measuring chamber
and disconnect the cable.
CAUTION
BEFORE REMOVING THE SENSOR, be
absolutely certain the process pressure is
reduced to 0 psig and the process temper
ature is at a safe level.
step 5
NOTE
If you have a dual input analyzer, you can reapply
power at this point. The initial reading from the
other sensor will be momentarily zero. After about
60 seconds the reading will reach its final value.
3. Using a small Phillips screwdriver, remove the two
screws holding the top flange of the sensor to the
body.
4. Using a slight back and forth twisting motion careful
ly pull the flange from the sensor body. You are
pulling against a single Oring seal. Don’t pull too
hard.
step 6
5. Using your thumb and forefinger, remove the
lamp/LED circuit board from the sensor.
6. Insert the replacement board in the sensor and push
the socket on the replacement board into the mating
pins in the sensor.
7. Place the desiccant package in the sensor body.
8. Orient the flange so that the screw holes line up with
the holes in the sensor body. Push the flange back
on the sensor body and replace the screws. Don’t let
wires push on lamp board. It may be necessary to
turn the flange a small amount until the holes line up.
46
FIGURE 83. Replacing the Lamp/LED Board
46
MODEL CLARITY II
SECTION 8.0
MAINTENANCE
9. Place the sensor in the calibration cup and reconnect the cable.
10. Calibrate the sensor using either slope or standard calibration (Section 6.2 or 6.3). Do not use grab calibra
tion. Failure to calibrate the sensor may reduce the life of the sensor. See Sections 8.2.5 and 8.2.6.
8.3 DEBUBBLER AND MEASURING CHAMBER
8.3.1 Cleaning the debubbler and measuring chamber
1. Turn off the sample supply to the debubbler.
WARNING
BEFORE DISCONNECTING THE SAMPLE AND
DRAIN LINES OR REMOVING THE SENSOR, be
absolutely certain the process pressure is reduced to
0 psig and the process temperature is at a safe level.
2. Remove the sensor and put it in a safe place. The calibration cup is a good place to store the sensor.
3. Loosen the small drain plug in the base plug and allow the sample in the debubbler to drain out. See Figure
84. Replace the drain plug.
4. Unscrew the upper and lower caps. Be careful not to lose the Orings.
5. Use a stream of water, a brush, or a rag to flush and clean out the inside of the debubbler and measuring
chamber.
6. Inspect the Orings for signs of damage and replace if necessary. The part number for the Oring (one each)
is 9550316.
7. Replace the upper and lower caps.
8. Replace the sensor.
8.3.2 Cleaning the orifice
1. Turn off the sample to the debubbler.
2. Disconnect the drain line. Unscrew the drain fitting from the orifice; then unscrew the orifice from the debub
bler body. See Figure 84.
3. Use a stream of water to flush out any residue accumulated in the orifice. Direct the stream of water counter
to the normal flow through the orifice.
4. If the material plugging the orifice cannot be removed with flushing, use a toothpick or a stiff wire to push out
the obstruction. Push counter to the normal flow through the orifice.
5. Reinstall the orifice and reconnect the drain line. Turn on the sample flow.
6. If the blockage cannot be removed or the orifice is damaged during cleaning, replace the orifice (PN 3394700).
47
MODEL CLARITY II
SECTION 8.0
MAINTENANCE
FIGURE 84. Molded Debubbler Assembly
8.4 LIST OF REPLACEMENT PARTS
LOCATION IN
FIGURE 84
—
—
—
—
1
2
3
4
5
6
7
8
9
not shown
48
DESCRIPTION
Replacement lamp board assembly, USEPAcompliant sensor
Replacement lamp board assembly, ISOcompliant sensor
Replacement sensor, USEPAcompliant
Replacement sensor, ISOcompliant
Debubbler housing
Upper cap for debubbler
Lower cap for debubbler
Sensor nut
Pipe plug, 1/4 inch MNPT ( 2 places)
Orifice assembly
Sample inlet elbow, 1/4 in compression fitting x 1/4 in MNPT
Sample drain elbow, 3/8 in barb x 1/4 in MNPT
Oring, one each, for upper and lower caps
Oring, one each, for sensor
PART
NUMBER
109010009EPA
109010010ISO
801080002EPA
801080003ISO
3401500
3401400
3401401
3401402
3000854
3394700
9321010
9322036
9550322
9550145
MODEL CLARITY II
SECTION 9.0
TROUBLESHOOTING
SECTION 9.0
TROUBLESHOOTING
9.1 OVERVIEW
The Model 1056 analyzer used in the Clarity II turbidimeter continuously monitors itself and the sensor for prob
lems. When the analyzer detects a problem, the word fault or warning followed by s appears in the display alter
nately with the measurement. If alarm 3 was configured as a fault alarm and a fault has occurred, the relay will
energize. The outputs do not change during a fault or warning condition. They continue to reflect the measured
turbidity or TSS value.
To read fault and warning messages, go to the main display and press s. The analyzer will
automatically scroll through the messages and will continue to scroll through the mes
sages for two minutes. After two minutes the display will return to the default screen.
To stop the automatic scrolling and return to the main display, press EXIT.
Error messages are prefaced by the word fault or warning.
Faults are conditions requiring immediate attention from the user. Measurements made by the analyzer should
be regarded as being in error.
Warnings are less serious than faults. A warning signifies the existence of a condition requiring attention. The
instrument remains usable.
WARNING
Explosion hazard. Do not disconnect equipment when
a flammable or combustible atmosphere is present.
9.2 TROUBLESHOOTING USING FAULT CODES
Fault message
SN Lamp/LED Failure
EEPROM Failure
Factory Failure
Explanation
Lamp or LED is burned out
Cannot save data to nonvolatile memory
Needs factory calibration
Section
9.2.1
9.2.2
9.2.3
Warning message
SN Need Cal
Explanation
Lamp intensity is weak but can be improved by
calibrating
Weak lamp, replace as soon as possible
Poor sensor cable connection or unusual ambient light
condition affecting sensor or sensor not immersed.
Section
9.2.4
SN Weak Lamp
SN Warning
9.2.5
9.2.6
SN identifies the sensor affected. S1 is sensor 1; S2 is sensor 2.
9.2.1 Lamp/LED Failure
The light source in a Clarity II turbidity sensor can be either a tungsten filament lamp or an LED. USEPAcompli
ant sensors use a tungsten lamp. ISOcompliant sensors use an LED. A photodiode inside the sensor continuously
monitors the intensity of the light source. The source intensity measurement is used to correct for source drift,
which allows the sensor to operate for longer periods without calibration. If the signal from the photodiode drops
below a certain value, the analyzer assumes the light source has either failed completely or the intensity is too low
to be useful. At this point the analyzer displays the Lamp Failure message.
Replace the lamp or LED board. See Section 7.2.2.
After replacing the lamp board, be sure to recalibrate the sensor using either slope or standard calibration. See
Section 6.2 or 6.3. Recalibration is necessary to reset the lamp power supply. Grab calibration will not reset the
power supply and may result in significantly reduced lamp life.
49
MODEL CLARITY II
9.2.2
SECTION 9.0
TROUBLESHOOTING
EEPROM Failure
EEPROM failure means the analyzer is unable to store data in the nonvolatile memory. Thus, if power is lost then
restored, all configurations and calibrations will be lost. Call the factory for assistance. The analyzer will probably
need to be replaced.
9.2.3
Factory Failure
Factory failure means the factory calibrations have been corrupted. Call the factory for assistance. The analyzer
will probably need to be replaced.
9.2.4
Need Cal
The Clarity II sensor contains two photodiodes. One measures the intensity of the light scattered by the sample.
The other measures the intensity of the lamp. Because turbidity is proportional to the intensity of light falling on the
sample photodiode, any reduction of the lamp intensity will be measured as a decrease in turbidity even though
the true turbidity remained constant. The analyzer uses the lamp intensity measurement to correct for changes in
apparent turbidity caused by reduction of lamp intensity. However, if the lamp intensity gets too low, the correction
may not be valid. At this point the analyzer displays the Need Cal warning. Calibrating will cause the analyzer to
increase the current supplied to the lamp, thus increasing the lamp intensity.
A. Calibrate the sensor using slope (Section 6.2), standard (Section 6.3), or lamp calibration (Section 6.5). Using
slope or standard calibration is strongly recommended. Use lamp calibration ONLY if a turbidity standard is not
available.
B. If a replacement lamp board is not available, order one as soon as possible.
9.2.5
Weak Lamp
The Weak Lamp warning appears when lamp intensity is low and the current being supplied to the lamp (see
Section 8.2.5) has been increased above a level likely to significantly reduce lamp life.
Replace the lamp board as soon as possible. After you replace the lamp, recalibrate the sensor using either slope
or standard calibration. See Section 6.2 or 6.3. Recalibration is necessary to reset the lamp power supply. Grab
calibration will not resent the power supply. Failure to recalibrate using slope or standard calibration may signifi
cantly reduce lamp life.
9.2.6
SN Warning
“SN Warning” will be displayed on the instrument to communicate and unusual but nonfatal condition that may
require checking and adjustments. Check three things.
A. Check the sensor/cable connection. Confirm that the swivel nut on the cable is in the locked position
on the sensor. Note: Once the plastic threaded swivel nut is engaged with the sensor threads, rotate
the swivel 3/4 turn to lock the cable to the sensor.
B. Confirm that sample water is flowing out of the debubbler drain outlet. This ensures that the sensor is
immersed in sample water.
C. Ensure that the sensor is not exposed to high ambient light sources (such as direct sunlight).
IMPORTANT NOTE:
When using EPA/incandescent sensors (PN 801080002EPA):
n
DO NOT power up the instrument without the sensor connected
n
DO NOT disconnect and reconnect a sensor while an analyzer is powered
If this is inconvenient or cannot be avoided:
5. Cycle power to the instrument after connecting the sensor or..
6. Perform a Slope Calibration or Standard Calibration routine after connecting the sensor
Following these guidelines will extend the life of the incandescent lamp and avoid premature warnings and faults
due to reduced lamp life.
50
MODEL CLARITY II
SECTION 9.0
TROUBLESHOOTING
9.3 TROUBLESHOOTING CALIBRATION PROBLEMS
Once the user has completed the calibration sequence, the analyzer verifies that the calibration meets certain
requirements. If the calibration is valid, the analyzer displays the calibration complete screen and updates the cal
ibration. If the calibration does not meet requirements, the calibration error screen appears. The analyzer retains
the original calibration.
Calibration method
Section
Userprepared standard (Section 6.2)
9.3.1
Commercial standard (Section 6.3)
9.3.2
Grab sample (Section 6.4)
9.3.3
9.3.1 Calibration ErrorUserPrepared Standard (Section 6.2)
A. For best results calibrate using freshly prepared 20.0 NTU standard. Use the procedure in Section 6.3.2.
B. Has the stock 4000 NTU standard exceeded its expiration date?
C. Is the turbidity of the dilution water less than 0.5 NTU? If you are using bottled distilled or deionized water,
open a fresh bottle and repeat the calibration.
D. Are the lamp and detector windows clean? See Section 7.2.1.
E. Is the sensor securely seated in calibration cup with no light leaking in? Putting a dark cloth over the sensor
and calibration cup and removing it should have no effect on the reading. Are both the lamp and photodiode
windows completely submerged in the standard?
F.
Was the correct turbidity value entered in the analyzer?
9.3.2 Calibration ErrorCommercial Standard (Section 6.3)
A. For best results calibrate using 20.0 NTU standard.
B. Has the calibration standard exceeded its expiration date?
C. Are the lamp and detector windows clean? See Section 7.2.1.
D. Is the sensor is securely seated in calibration cup with no light leaking in? Putting a dark cloth over the sen
sor and calibration cup and removing it should have no effect on the reading. Are both the lamp and photodi
ode windows completely submerged in the standard?
E. Was the correct turbidity value entered in the analyzer?
9.3.3 Calibration ErrorGrab Sample (Section 6.4)
A. Was the referee instrument used to measure the grab sample properly calibrated?
B. Was the process turbidity reading stable when the grab sample was taken? Do not attempt a grab sample cal
ibration when turbidity readings are rapidly changing.
C. Is the sensor securely seated in the measuring chamber with no light leaking in? Putting a dark cloth over the
sensor and measuring chamber and removing it should have no effect on the reading.
D. Is the sensor clean? See Section 7.2.1.
E. Was the correct turbidity value entered in the analyzer?
51
MODEL CLARITY II
SECTION 9.0
TROUBLESHOOTING
9.4 TROUBLESHOOTING OTHER PROBLEMS
Problem
Section
Readings are erratic
9.4.1
Readings drift
9.4.2
Analyzer responds too slowly to changes in turbidity
9.4.3
Flow is too low
9.4.4
Readings are lower or higher than expected
9.4.5
Current output is too low
9.4.6
Alarm relays do not operate when setpoint is exceeded
9.4.7
Display is unreadabletoo faint or all pixels dark
9.4.8
9.4.1 Readings are erratic
Erratic readings are usually caused by air bubbles drifting through the measurement zone of the sensor. Air bub
bles reflect light onto the detector and cause spikes in the turbidity reading. A debubbling chamber helps remove
large bubbles. An orifice in the outlet of the debubbler helps eliminate outgassing by putting back pressure on the
debubbler. Outgassing can occur when the pressure of the sample is reduced or when a cold sample warms up.
A bubble rejection filter in the analyzer software also helps reduce the effect of bubbles.
A. Be sure the bubble rejection filter is on and increase the signal averaging time. See Section 5.5.
B. If the inlet pressure is high enough, increase the back pressure on the debubbler using a valve or a valved
rotameter (PN 2410300) installed in the outlet of the debubbler. Do not exceed 30 psig (308 kPa abs).
Increasing the back pressure reduces the sample flow and increases the system response time. If the inlet
pressure is too low, increasing the back pressure might not be feasible.
C. If bubbles persist, increase the back pressure and use a sample pump to increase the inlet pressure.
9.4.2 Readings drift
Gradual downward drift in readings is caused by dirt accumulating on the lamp or detector windows. The dirt
reduces the amount of light entering the measuring zone in the sample and blocks scattered light from reaching
the detector. Upward drift is usually caused by bubbles adhering to the lamp or detector windows. The bubbles,
which act like lenses, direct light onto the detector and increase the apparent turbidity reading. Once the bubbles
get large enough, they break away from the surface of the detector and the turbidity reading drops.
A. If downward drift is occurring, inspect the sensor windows for cleanliness. See Section 7.2.1 for cleaning
instructions.
B. If upward drift is occurring, remove the sensor completely from the debubbler and then replace it. If readings
drop back to normal or expected values, then the upward drift was probably caused by bubbles accumulating
on the sensor. (Removing the sensor from the debubbler causes the air bubbles to break.) To reduce bubble
accumulation, increase the back pressure on the debubbler using a valve or valved rotameter (PN 2410300)
installed in the outlet of the debubbler. Do not exceed 30 psig (308 kPa abs). Increasing the back pressure
reduces the sample flow and increases the system response time. If the inlet pressure is too low, increasing
the back pressure might not be feasible.
C. If bubbles persist, increase the back pressure and use a sample pump to increase the inlet pressure.
52
MODEL CLARITY II
SECTION 9.0
TROUBLESHOOTING
9.4.3 Analyzer responds too slowly to changes in turbidity
Response time is primarily a function of sample flow rate, distance between the sample point and analyzer, and
the diameter of the sample tubing. Because the debubbler has a flow restrictor on the outlet to increase back pres
sure, sample flow rate is primarily determined by the inlet pressure.
A. If possible, increase the inlet pressure.
B. If increasing the inlet pressure is not feasible, move the sensor closer to the sample point.
9.4.4 Flow is too low
The debubbler is fitted with a 0.040 inch (1 mm) diameter orifice in the outlet. The orifice puts back pressure on
the debubbler, which helps reduce outgassing. If the inlet pressure is about 3.5 psig (125 kPa abs), the flow
through the debubbler will be about 250 mL/min. The response time to a step change at 250 mL/min is about sss
minutes. If the flow is too low, the response time may become excessive. The only way to improve the response
time is to reduce the back pressure or to increase the inlet pressure.
A. To eliminate back pressure, remove the orifice from the debubbler. See Section 7.3.
B. If removing the orifice causes outgassing — the symptom of outgassing is an upward drift in apparent turbid
ity — increase the back pressure by a small amount. Use a valve or a valved rotameter (PN 2410300) in the
debubbler outlet. Do not exceed 30 psig (308 kPa abs).
C. If outgassing continues to persist, increase the back pressure. To maintain flow, use a pump to increase the
inlet pressure.
9.4.5 Readings are lower or higher than expected
A. Is the instrument to which readings are being compared properly calibrated?
B. Are samples being tested immediately after sampling? If samples are allowed to sit too long before testing, the
turbidity may change.
C. Are the measurement chamber and debubbler clean?. Sample flow may be stirring up solids that have previ
ously settled out in the debubbler and measurement chamber, increasing the apparent turbidity. See Section
7.3.1 for cleaning procedure.
9.4.6 Analog current is too low
Load resistance is too high. Maximum load is 600 Ω.
9.4.7 Alarm relays do not operate when setpoint is exceeded
A. Is the alarm board is in place and properly seated?
B. Is the alarm logic (high/low) and dead band correct?
C. Has the setpoint has been properly entered?
9.4.8 Display is unreadable — too faint or all pixels dark.
While holding down the MENU key, press s or t until the display has the correct contrast.
53
MODEL CLARITY II
SECTION 8.0
TROUBLESHOOTING
This page left blank intentionally
54
MODEL CLARITY II
SECTION 10.0
RETURN OF MATERIAL
SECTION 10.0
RETURN OF MATERIAL
10.1 GENERAL.
To expedite the repair and return of instruments, proper communication between the customer and the factory is
important. Before returning a product for repair, call 19497578500 for a Return Materials Authorization (RMA)
number.
10.2 WARRANTY REPAIR.
The following is the procedure for returning instruments still under warranty:
1.
Call Rosemount Analytical for authorization.
2.
To verify warranty, supply the factory sales order number or the original purchase order number. In the case
of individual parts or subassemblies, the serial number on the unit must be supplied.
3.
Carefully package the materials and enclose your “Letter of Transmittal” (see Warranty). If possible, pack the
materials in the same manner as they were received.
4.
Send the package prepaid to:
Emerson Process Management
Liquid Division
2400 Barranca Parkway
Irvine, CA 92606
Attn: Factory Repair
RMA No. ____________
Mark the package: Returned for Repair
Model No. ____
10.3 NONWARRANTY REPAIR.
The following is the procedure for returning for repair instruments that are no longer under warranty:
1.
Call Rosemount Analytical for authorization.
2.
Supply the purchase order number, and make sure to provide the name and telephone number of the individ
ual to be contacted should additional information be needed.
3.
Do Steps 3 and 4 of Section 9.2.
NOTE
Consult the factory for additional information regarding service or repair.
55
MODEL CLARITY II
APPENDIX
APPENDIX
This procedure describes how to verify linearity between turbidity and TSS.
1. Collect a sample of the process liquidyou may need 10 L or more if you use the Clarity II for measuring tur
bidity. If you use a laboratory turbidimeter, you will need less volume. The Clarity II requires about 500 mL
for the measurement; laboratory turbidimeters require 50 mL or less. Verify that the turbidity of the sample is
less than 200 NTU. Store the in a clean bottle.
2. Filter a portion of the sample to obtain at least 5 L of dilution liquid. The filtrate is needed to dilute the sam
ple in subsequent steps. Verify that the turbidity of the dilution water is low. If filtering the sample is impracti
cal, use deionized water for dilution.
3. Measure the total suspended solids (TSS) in the sample obtained in step 1. Thoroughly mix the sample
before withdrawing liquid. A magnetic stirrer is best, but inverting the sample repeatedly for about a minute
works, too. Avoid violent shaking or mixing. Refer to any standard reference work on water and wastewater
testing for the procedure for determining TSS.
4. Dilute the sample from step 1, by a factor of 0.9, 0.7, 0.5, 0.3, and 0.1. See the table for recommended vol
umes. Measure TSS and turbidity for each dilution. For lower TSS values, use a larger volume of sample.
Dilution
Volume of
Final
Volume for
Volume for
factor
stock, mL
volume, mL
Clarity II, mL
TSS, mL
1.00
500
50 250
0.9
900
1000
500
50 250
0.7
700
1000
500
50 250
0.5
500
1000
500
50 250
0.3
300
1000
500
50 250
0.1
100
1000
500
50 250
5. Plot the data obtained in step 4, with turbidity on the yaxis and TSS on the xaxis. Fit the best straight line
to the data.
6. Locate two points (P1 and P2) on the line separated as much as possible. Read the ppm and NTU value for
each point and enter these into the analyzer. See Section 6.5.2.
56
WARRANTY
Seller warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufactured
or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expira
tion of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen
(18) months from the date of shipment by Seller, whichever period expires first. Consumables, such as glass electrodes, mem
branes, liquid junctions, electrolyte, orings, catalytic beads, etc., and Services are warranted for a period of 90 days
from the date of shipment or provision.
Products purchased by Seller from a third party for resale to Buyer ("Resale Products") shall carry only the warranty extended by
the original manufacturer. Buyer agrees that Seller has no liability for Resale Products beyond making a reasonable commercial
effort to arrange for procurement and shipping of the Resale Products.
If Buyer discovers any warranty defects and notifies Seller thereof in writing during the applicable warranty period, Seller shall, at
its option, promptly correct any errors that are found by Seller in the firmware or Services, or repair or replace F.O.B. point of man
ufacture that portion of the Goods or firmware found by Seller to be defective, or refund the purchase price of the defective por
tion of the Goods/Services.
All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuit
able environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any other
cause not the fault of Seller are not covered by this limited warranty, and shall be at Buyer's expense. Seller shall not be obli
gated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by
an authorized Seller representative. All costs of dismantling, reinstallation and freight and the time and expenses of Seller's per
sonnel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Seller.
Goods repaired and parts replaced during the warranty period shall be in warranty for the remainder of the original warranty peri
od or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Seller and can be amended only
in a writing signed by an authorized representative of Seller. Except as otherwise expressly provided in the Agreement, THERE
ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESS OR IMPLIED, AS TO MERCHANTABILITY, FIT
NESS FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES.
RETURN OF MATERIAL
Material returned for repair, whether in or out of warranty, should be shipped prepaid to:
Emerson Process Management
Liquid Division
2400 Barranca Parkway
Irvine, CA 92606
The shipping container should be marked:
Return for Repair
Model _______________________________
The returned material should be accompanied by a letter of transmittal which should include the following information (make a
copy of the "Return of Materials Request" found on the last page of the Manual and provide the following thereon):
1.
2.
3.
4.
5.
Location type of service, and length of time of service of the device.
Description of the faulty operation of the device and the circumstances of the failure.
Name and telephone number of the person to contact if there are questions about the returned material.
Statement as to whether warranty or nonwarranty service is requested.
Complete shipping instructions for return of the material.
Adherence to these procedures will expedite handling of the returned material and will prevent unnecessary additional charges
for inspection and testing to determine the problem with the device.
If the material is returned for outofwarranty repairs, a purchase order for repairs should be enclosed.
The right people,
the right answers,
right now.
ONLINE ORDERING NOW AVAILABLE ON OUR WEB SITE
http://www.rosemountanalytical.com
Specifications subject to change without notice.
8
Credit Cards for U.S. Purchases Only.
Emerson Process Management
2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 7578500
Fax: (949) 4747250
http://www.rosemountanalytical.com
© Rosemount Analytical Inc. 2014