Servomex Servotough Spectrascan (2400) Gas Analyzer User Manual
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
PROCESS ANALYZERS
SERVOTOUGH SpectraScan (2400)
Gas Analyzer
User Manual
Part Number:
Revision:
Language:
02400001A
0
US English
The SpectraScan Analyzer is manufactured by Precisive LLC of Boston , MA, USA who
assume responsibility for the product performance with regard to software and set up
using appropriate gas mixtures for calibration. The methodology is patent protected
and is the Intellectual Property of Precisive LLC.
SERVOTOUGH SpectraScan User Manual
TABLE OF CONTENTS
1
ABOUT THIS MANUAL ....................................................................................... 6
1.0
1.1
1.1.1
1.1.2
1.1.3
Introduction .......................................................................................................... 6
Symbols and Notices Used in this Manual .......................................................... 7
Hazard Notices .................................................................................................... 7
Hazard Labels Statement .................................................................................... 7
IEC and ISO Symbols.......................................................................................... 8
2
SYSTEM OVERVIEW ......................................................................................... 9
2.0
2.1
Introduction to the Analyzer ................................................................................. 9
Measurement Principle ........................................................................................ 9
3
SYSTEM SAFETY ............................................................................................. 12
3.0
3.1
3.2
3.3
3.4
3.4.1
3.4.2
3.4.3
3.4.4
3.4.5
3.4.6
General Precautions .......................................................................................... 13
WARNINGS AND CAUTIONS .......................................................................... 15
Suitability of Application..................................................................................... 17
Operator Training .............................................................................................. 17
Analyzer Hazards .............................................................................................. 17
Gases ................................................................................................................ 17
Heated Equipment ............................................................................................. 18
Over Pressurizing .............................................................................................. 18
Electrical Safety ................................................................................................. 19
Electrostatic Discharge ...................................................................................... 19
Ergonomic Considerations ................................................................................ 20
4
UNPACKING THE ANALYZER ......................................................................... 21
4.0
4.1
4.2
Component List ................................................................................................. 21
Removing the Analyzer from the Shipping Container ....................................... 22
Inspection .......................................................................................................... 22
5
MOUNTING CONSIDERATIONS ..................................................................... 23
5.0
5.0.1
5.0.2
5.0.3
5.0.4
5.0.5
5.1
Environmental Requirements ............................................................................ 23
Dimensions, Weight and Ventilation.................................................................. 24
Operating Temperature, Humidity and Pressure .............................................. 25
Combustion Hazard ........................................................................................... 26
Power Requirements ......................................................................................... 26
Fittings Requirements........................................................................................ 26
Mounting Procedure .......................................................................................... 26
6
ELECTRICAL CONNECTIONS ......................................................................... 29
6.0
6.0.1
6.1
6.2
6.2.1
6.2.2
6.3
6.4
I/O Panel Layout ................................................................................................ 30
Component Layout ............................................................................................ 32
Electrical Power Requirements ......................................................................... 34
Wiring................................................................................................................. 35
Power Wiring: 24 VDC....................................................................................... 36
Power Wiring: 120-240 VAC.............................................................................. 38
Communication Connection .............................................................................. 40
Spectrascan Networking and Software Configuration....................................... 42
02400001A/0
i
SERVOTOUGH SpectraScan User Manual
6.4.1
6.4.2
6.4.3
6.4.4
6.4.5
Ethernet and TCP/IP ......................................................................................... 42
Finding the IP address of your instrument......................................................... 42
Setting a static IP address ................................................................................. 43
RS-422 connection ............................................................................................ 46
Modbus register ................................................................................................. 48
7
SAMPLE GAS CONNECTIONS ........................................................................ 49
7.0
7.0.1
7.0.2
7.0.3
7.0.4
7.0.5
7.0.6
7.1
7.1.1
Sample Line Connections ................................................................................. 50
Sampling Ports .................................................................................................. 50
Preparing Sample Gas Lines ............................................................................ 50
Sampling Gas from a Pressurized Source ........................................................ 51
Cleanliness of Sample Stream .......................................................................... 51
Stream Switching Using External Solenoid Valves ........................................... 52
Procedure for Connecting the Sample Lines..................................................... 54
Enclosure Purge Connections ........................................................................... 55
Purging of Enclosure Compartment .................................................................. 55
8
INSTALLATION CHECKLIST ............................................................................ 57
9
VERIFICATION ................................................................................................. 59
9.0
9.0.1
9.0.2
9.0.3
9.1
Procedure .......................................................................................................... 59
Setup ................................................................................................................. 59
Zeroing............................................................................................................... 60
Span Verification ............................................................................................... 60
Re-spanning ...................................................................................................... 60
10
MAINTENANCE ................................................................................................ 61
10.0
Light Source Assembly Replacement Procedure .............................................. 62
11
TROUBLESHOOTING ...................................................................................... 67
12
APPENDIX A: PRODUCT SPECIFICATIONS .................................................. 69
12.0
12.1
12.2
12.3
12.4
12.5
12.6
Measurement Specifications ............................................................................. 69
Physical Specifications ...................................................................................... 70
Communications Specification .......................................................................... 70
Sampling System Specifications ....................................................................... 71
Electrical Specifications ..................................................................................... 72
Environmental Specifications ............................................................................ 72
Manual Revision History .................................................................................... 72
13
INDEX ................................................................................................................ 73
14
CONTACT DETAILS ......................................................................................... 74
02400001A/0
ii
SERVOTOUGH SpectraScan User Manual
List of Figures
Figure 1 - The Servomex SpectraScan Gas Analyzer ............................................................. 23
Figure 2 - Hole Dimensions of the Servomex SpectraScan Gas Analyzer.............................. 24
Figure 3 - I/O panel of the Servomex SpectraScan Gas Analyzer .......................................... 30
Figure 4 - Analyzer with Cover Open ....................................................................................... 31
Figure 5 - Interior of the Servomex SpectraScan Gas Analyzer .............................................. 32
Figure 6 - SpectraScan Interior Diagram ................................................................................. 33
Figure 7 - Example Rating Label ............................................................................................. 34
Figure 8 - Wiring the SpectraScan with 24 VDC ...................................................................... 36
Figure 9 - Wiring the SpectraScan with 120-240Vac ............................................................... 38
Figure 10 - Connecting the Ethernet TCP/IP communication cable ........................................ 40
Figure 11 - Location of the serial number tag .......................................................................... 42
Figure 12 - Location of debug RS-232 connector on MPU ...................................................... 44
Figure 13 - Sample session in which a static IP address is set using PuTTY to communicate
with the MPU..................................................................................................................... 46
Figure 14 - Identification of RS-422 signals in instrument ....................................................... 47
Figure 15 - Location of RS-422 connector on system board ................................................... 47
Figure 17 – Typical Gas Connection Configuration ................................................................. 49
Figure 18 - Recommended Connections for Inlet Stream Switching ....................................... 53
Figure 19 - Accessing the Light Source ................................................................................... 63
Figure 20 - The Light Source Circuit Board ............................................................................. 63
Figure 22 - Mounting Hole for Light Source ............................................................................. 64
Figure 21 - The Light Source Assembly................................................................................... 64
Figure 23 - Dimensions of SpectraScan .................................................................................. 70
02400001A/0
iii
SERVOTOUGH SpectraScan User Manual
List of Tables
Table 1 - Definition of Symbols Found on the Analyzer............................................................. 8
Table 2 - Hazardous Location Ratings..................................................................................... 12
Table 3 - Standard Fittings Used on the Servomex SpectraScan Gas Analyzer .................... 26
Table 4 - Wire Types ................................................................................................................ 35
Table 5 - SpectraScan Modbus register .................................................................................. 48
Table 6 - Diagnostic LED Indications ....................................................................................... 59
Table 7 - Troubleshooting ........................................................................................................ 67
Table 8 – Measurement Specifications .................................................................................... 69
Table 9 – Physical Specifications ............................................................................................ 70
Table 10 - Communications Specifications .............................................................................. 70
Table 11 - Sampling System Specifications ............................................................................ 71
Table 12 - Response Time Versus Gas Flow Rate ................................................................. 71
Table 13 - Electrical Specifications .......................................................................................... 72
Table 14 - Environmental Specifications.................................................................................. 72
02400001A/0
iv
SERVOTOUGH SpectraScan User Manual
This page intentionally left blank.
02400001A/0
v
SERVOTOUGH SpectraScan User Manual
1 ABOUT THIS MANUAL
1.0 Introduction
This manual provides information about the Servomex SpectraScan Gas Analyzer. TCP/IP enabled,
the Servomex SpectraScan Gas Analyzer can monitor process gases remotely, unattended for
extended periods of time.
This manual presents all safety considerations of the Servomex SpectraScan Gas Analyzer first,
and then provides instructions for unpacking, mounting and connecting the unit. An installation
checklist ensures all the required tasks are performed. The verification test confirms span accuracy
upon installation, and includes the zeroing procedure which may be repeated periodically. Short
maintenance and troubleshooting sections are included in this manual. Product specifications are
given in the appendix at the end of this manual.
Before installing or operating your Servomex SpectraScan Gas Analyzer, carefully read and
familiarize yourself with System Safety. In addition, observe and obey all Hazard Notices and
Symbols provided for your safety throughout this manual.
02400001A/0
NOTES
Read this manual in its entirety before attempting to perform any operations
on the Servomex SpectraScan System.
6
SERVOTOUGH SpectraScan User Manual
1.1 Symbols and Notices Used in this Manual
This manual contains attention symbols and notices which should be observed to ensure personal
safety, as well as to protect the product and the connected equipment. Hazard notices are
accompanied with text that identifies the hazard, advises you to avoid the hazard, and/or the
probable consequences of NOT avoiding the hazard. Where applicable, within the notice box or
near the paragraph text, the appropriate IEC or ISO symbol is used. Attention notices provide useful
information pertinent to the operation of the analyzer.
The following notices are displayed throughout Servomex documentation:
1.1.1
Hazard Notices
Read this manual and ensure that you fully understand its content before you attempt to install, use
or maintain the analyzer. Important safety information is highlighted in this manual as WARNINGS
and CAUTIONS and NOTES, which are used as follows;
WARNING
Warnings highlight specific hazards which, if not taken into account, may result
in personal injury or death.
CAUTION
Cautions highlight hazards which, if not taken into account, can result in damage to the analyser
or to other equipment or property.
1.1.2
NOTES
This highlights information which it is useful for you to be aware of (for
example, specific operating conditions, and so on).
Hazard Labels Statement
Hazard labels indicate possible hazardous areas and identify the potential hazard on the analyzer.
Hazard labels are applied to the access areas of the analyzer to caution the following:
Exposure to hazardous chemicals (gases, process materials or by-products,
solvents/solutions) or physical conditions (heat or excessive cold) that may result in injuries
Exposure to electrical energy that may cause electric shock or death
Exposure to high temperatures that may cause injuries
See Table 1 IEC and ISO Symbols on page 8 for definition of symbols.
02400001A/0
7
SERVOTOUGH SpectraScan User Manual
1.1.3
IEC and ISO Symbols
The following symbols may be found on the Servomex SpectraScan Gas Analyzer and in the
documentation:
Definition of Symbols Found on the Analyzer and in this Manual
|
On (Supply)
Off (Supply)
Protective Earth
(Ground)
Protective Conductor
Terminal
IEC 60417, No. 5007
IEC 60417, No. 5008
IEC 60417, No. 5017
IEC 60417, No. 5019
Caution, hot surface
Equipotentiality
Frame or Chassis
Terminal
ISO 3684, No. 6043
IEC 60417, No. 5021
IEC 6041, No. 5020
Double or Reinforced
Insulation
IEC 6041, No. 5172
3
Direct Current
Alternating Current
Both Direct and
Alternating Current
IEC 60417, No. 5031
IEC 60417, No. 5032
IEC 60417, No. 5033
Three-phase
Alternating Current
Caution, refer to
Caution, risk of
electric shock
ISO 3864, No.B.3.6
accompanying
documents
Out position of a bistable push control
In position of a bistable push control
ISO 3864, No.B.3.1
IEC 60417, No 5269
IEC 60417, No 5268
Table 1 - Definition of Symbols Found on the Analyzer
02400001A/0
8
SERVOTOUGH SpectraScan User Manual
2 SYSTEM OVERVIEW
This chapter introduces the Servomex SpectraScan Gas Analyzer and explains its basic principle of
operation.
2.0 Introduction to the Analyzer
The Servomex SpectraScan Gas Analyzer is a non-contact, light absorption based gas analyzer
capable of ppm to percent level concentration monitoring of multiple gas compounds. The system
consists of a light spectrometer, a flow-through sample cell, a single-element photo-detector and the
supporting electronics. The spectrometer uses a unique tunable Fabry-Perot assembly that
provides wavelength scanning with high optical throughput. An advanced spectral processing
algorithm computed in the embedded electronics provides highly accurate and robust quantitative
measurements.
The analyzer is configured and calibrated for a specific wavelength analysis region(s) depending on
the application for which it is intended. It is designed to be a dedicated on-line monitoring system
that is “plug and play,” not requiring any on-site configuration or calibration. The specific system
configuration and calibration is denoted by the application code “-xyz” that can be found in the
Calibration Verification Sheet delivered with the analyzer.
Example applications of the Servomex SpectraScan Gas Analyzer include:
Hydrocarbon gas composition monitoring
Petrochemical process monitoring
Specialty and chemical gas process and blending monitoring
Catalysis and combustion process monitoring
2.1 Measurement Principle
When a gas sample is introduced in the gas cell, the light radiation provided by a broadband light
source is partially absorbed by the gas species present. The light absorption occurs at specific
frequencies and magnitudes depending on the gas compound and the concentration of that
compound. The Servomex SpectraScan spectrometer module scans the wavelength and measures
the true absorption spectra and compares them with the pre-loaded calibration spectra. The onboard analysis algorithm computes the gas concentrations in real-time, which can then be output
through the MODBUS TCP/IP protocol.
In principal, the absorption spectrum of each compound is unique which acts as a “fingerprint” for
identification or speciation analysis. In addition, the magnitude of the absorption is a function of the
number of molecules of the gas. With a known path length, pressure and temperature, the
magnitudes of the absorption spectra are then used to compute volumetric concentrations. This
first principle based technique provides accurate and robust measurements with minimal span and
baseline drifts.
The analyzer employs an internal pressure transducer to measure the sample pressure in real-time
enabling pressure variation corrections. The flow cell is heated to a constant temperature (default
value is 60oC) with a sample preheat module to maintain both sample and optical sensor
temperature at a constant calibrated temperature, thereby ensuring measurement accuracy and
stability despite sample and environmental variations.
02400001A/0
9
SERVOTOUGH SpectraScan User Manual
02400001A/0
10
SERVOTOUGH SpectraScan User Manual
This page intentionally left blank.
02400001A/0
11
SERVOTOUGH SpectraScan User Manual
3 SYSTEM SAFETY
A primary concern of Servomex is to ensure a safe working environment for all users of our system.
The Servomex SpectraScan Gas Analyzer meets the following classifications;
Hazardous Location Classification:
Classification and Rating
Class 1, Division 2, Groups A,B,C and D, T4
Class 1, Zone 2, AEx nA nC nL IIC T4 Gc
Class 1, Zone 2, Ex nA nC nL IIC T4 Gc
II 3G Ex ic nA nC IIC T4 Gc
IP66
ITS12ATEX
Ambient Temperature Rating: -20 < Tamb < 50°C (-4 to 122°F)
Table 2 - Hazardous Location Ratings
It is essential that all users become aware of the hazards associated with the start-up, operation
and maintenance of this analyzer. Servomex SpectraScan users must observe all prescribed safety
guidelines and adopt good general work practices to adequately protect themselves, fellow workers
and the unit. Only trained and experienced users should start, operate and/or maintain the
Servomex SpectraScan Gas Analyzer.
WARNING
The following general safety precautions must be observed during all phases
of installation and operation of this analyzer. Failure to comply with these
precautions or with specific warnings elsewhere in this manual violates safety
standards of intended use of the unit and may impair the protection provided
by the analyzer, exposing the user to the risk of injury or death and possibly
causing damage to the analyzer. Servomex assumes no liability for the
customer’s failure to comply with these requirements.
02400001A/0
12
SERVOTOUGH SpectraScan User Manual
3.0 General Precautions
DO NOT SUBSTITUTE PARTS OR MODIFY THIS ANALYZER
Do not install substitute parts or perform any unauthorized modification to the analyzer. Return the
analyzer to Servomex for service and repair to ensure that all safety features are maintained.
SERVICE BY QUALIFIED PERSONNEL ONLY
There are no user serviceable components in the analyzer. Component replacement and internal
adjustments must be made by qualified Servomex representatives only.
GROUNDING THE PRODUCT
This product is grounded through the grounding conductor of the power cabling to the ground lug
inside the enclosure. This grounding is essential for safe operation. Upon loss of the protective
ground connection, all accessible conductive parts (including knobs and controls that may appear to
be insulated) can render an electrical shock. Dangerous voltages are contained within this
analyzer. All electrical fittings and cables must be of the type specified, and be in good condition,
and all electrical fittings must be properly connected and grounded. All cables entering the analyzer
(24 VDC or 120/240 VAC power and communication) must be separately jacketed as they enter the
analyzer to prevent mis-wiring.
symbol for ground
USING THE PROPER POWER CABLING
This product is designed to be hardwired and has a 24 VDC model and a 120/240 VAC model. Use
only power cabling that is in good condition and which meets the input power requirements
specified in this manual. The power cabling used, whether for 24 VDC operation or for 120-240
VAC operation, must use a 3 conductor cable that is jacketed separately from the communications
cable. The ground wire in this 3 conductor cable must be electrically connected to the earth ground
lug in the analyzer. See Section 5.0.4 Power Requirements on page 26 and Table 13 - Electrical
Specifications on page 72 for more information. Also see Table 4 - Wire Types on page 35.
USING THE PROPER POWER SOURCE
This product is configured at the factory to operate from a 24 VDC or 120-240 VAC power source,
depending on the model. See labelling on the analyzer to determine which power source to use.
Ensure that the power supplied to the analyzer is appropriate for the configuration of analyzer being
installed. See Section 5.0.4 Power Requirements on page 26 for more information.
02400001A/0
13
SERVOTOUGH SpectraScan User Manual
USE THE PROPER FUSE
Use only a fuse of the correct type, voltage rating, and current rating, as specified for your product.
See Table 13 - Electrical Specifications on page 72 and Figure 6 - SpectraScan Interior Diagram
on page 33 for more information.
DO NOT OPERATE IN EXPLOSIVE ATMOSPHERES
To avoid explosion, do not operate this product in an explosive environment unless it has been
specifically certified for such operation. If it has been certified for such operation, carefully follow
the guidelines set forth in the Material Safety Data Sheets for the explosive chemical, as well as the
product labelling on the Servomex SpectraScan Gas Analyzer.
HIGH VOLTAGE DANGER
Assume high voltage is present in the cable and in the analyzer whenever a power cord is
connected.
USE PROPER FITTINGS AND TIGHTENING PROCEDURES
All connection fittings must be consistent with Servomex SpectraScan Gas Analyzer specifications,
and compatible with the intended use of the unit. Assemble and tighten fittings according to
manufacturer's directions. See Section 5.0.5 Fittings Requirements on page 26 for more
information.
CHECK FOR LEAK-TIGHT FITTINGS
Carefully check all connections to ensure leak-tight installation.
OPERATE AT SAFE INLET PRESSURES
Never operate at pressures higher than the rated maximum and minimum pressure. See Section
3.4.3 Over Pressurizing on page 18 and Table 11 - Sampling System Specifications on page 71 for
more information.
KEEP THE UNIT FREE OF CONTAMINANTS
Do not allow contaminants to enter the unit before or during use. Contamination such as dust, dirt,
lint, glass chips, or metal chips may permanently damage the unit.
02400001A/0
14
SERVOTOUGH SpectraScan User Manual
ALLOW PROPER WARM-UP TIME
The unit’s gas cell is temperature-controlled and will only meet measurement accuracy
specifications when sufficient time is allowed for the unit to meet and stabilize to the designed
operating temperature. Warm-up time for the unit is approximately 30 minutes to 2 hours depending
on the ambient temperature. The analyzer temperature shall be stable (default value is 60oC) prior
to performing any measurement. The red “Warm-up” LED on the display extinguishes and the green
“Run” LED lights when the unit has reached operating temperature.
MAINTAIN INGRESS PROTECTION
The Servomex SpectraScan unit has been tested to an Ingress Protection level of IP66, Type 4. In
order to maintain this rating, a suitable 1” conduit connector must be used that will not compromise
the existing IP Rating.
3.1
WARNINGS AND CAUTIONS
This apparatus is suitable for use in Hazardous Locations as defined in Table 2 - Hazardous
Location Ratings on page 12, or unclassified locations.
WARNING
LOSS OF SYSTEM INTEGRITY: Exposure to some chemicals may degrade
the sealing properties.
WARNING
EXPLOSION HAZARD: Do not remove or replace fuses unless power has
been disconnected or the area is known to be free of ignitable concentrations
of flammable gases or vapours.
WARNING
EXPLOSION HAZARD: The ac and dc source must be located in an
unclassified area.
WARNING
SYSTEM SAFETY: This analyzer is typed as permanently connected
equipment and as such, must include a switch or circuit–breaker in the
installation. This switch must be supplied and installed by the customer (not
provided by Servomex. It must be suitably located and easily reached, and
must be marked as the disconnecting device for the equipment.
02400001A/0
15
SERVOTOUGH SpectraScan User Manual
WARNING
EXPLOSION HAZARD: Do not disconnect while the circuit is live unless the
area is known to be free of ignitable concentrations of flammable gases or
vapours.
WARNING
EXPLOSION HAZARD: Substitution of components may impair suitability for
class 1, division 2 and class 1, zone 2 installations.
WARNING
EXPLOSION HAZARD: Do not use screws that exceed the maximum screw
penetration depth in any threaded holes.
02400001A/0
16
SERVOTOUGH SpectraScan User Manual
3.2 Suitability of Application
It is the customer’s responsibility to determine the suitability of the Servomex SpectraScan Gas
Analyzer for a given purpose. Servomex makes no representation that potential hazards presented
here or elsewhere are the only hazards associated with this product and its start-up, operation or
maintenance.
3.3 Operator Training
All users should be fully trained in all applicable safety and emergency procedures. At a minimum:
Safe work practices and procedures
Electrical safety
Chemical safety and procedures for handling hazardous materials
Proper use of personal protective equipment (PPE)
3.4 Analyzer Hazards
The Servomex SpectraScan is designed to be safe. The following hazards are unavoidably present
when operating or servicing the Servomex SpectraScan Gas Analyzer. All personnel connected to
the operation, installation or maintenance of the Servomex SpectraScan Gas Analyzer must be
familiar with the hazards intrinsic to the analyzer’s environment:
3.4.1
Gases
Heated Equipment
Over Pressurizing
Electrical Safety
Electrostatic Discharge
Ergonomic Considerations
Gases
WARNING
Imminent Danger of Personal Injury or Death: The Servomex SpectraScan
Gas Analyzer is designed and can be configured to analyze some flammable
or explosive gas mixtures. Do not use this unit with flammable or explosive
gases unless it has been specifically certified for that application.
02400001A/0
NOTE
Consult the gas manufacturer’s Material Safety Data Sheets (MSDS) for
further details on the storage, use, and safe handling procedures applicable
for these gases.
17
SERVOTOUGH SpectraScan User Manual
3.4.2
Heated Equipment
The gas cell in the Servomex SpectraScan Gas Analyzer will be heated to a temperature that is
higher than the gas being sampled to avoid condensation on the optical windows in the gas cell.
WARNING
Burn Hazard: The gas cell, though insulated, may be extremely hot. Use
caution when handling the gas cell, sample lines or other heated equipment.
3.4.3
Over Pressurizing
CAUTION
Equipment Damage: Restricting the gas flow on either the intake or the exhaust side may
cause the gas cell to over pressurize, causing damage to the analyzer. Care should be taken to
remove any port caps or other flow barriers before connecting any sample line or switching on
the analyzer. Also make sure the exhaust routing is appropriate for your specific application.
When the gas is to be sampled from a pressurized source, such as a pressurized line, a pressure
relief mechanism should be employed so as not to force or restrict the flow of gas into the unit.
CAUTION
Equipment Damage: Pressure relief mechanism should be employed when sampling gas from
a source other than the ambient in order to avoid damage to the analyzer.
02400001A/0
18
SERVOTOUGH SpectraScan User Manual
3.4.4
Electrical Safety
The Servomex SpectraScan Gas Analyzer is powered by voltages that could pose a hazard to the
user. Before any maintenance work is performed, disconnect power from the analyzer and turn off
the gas supply. Follow standard electrical safe work practices to prevent injuries.
WARNING
Electrical Hazard, Personal Injury and Equipment Damage: Do NOT
perform maintenance on an energized or gas supplied analyzer. When
performing maintenance on the unit, it is necessary to disconnect the power
source from the unit, and turn off the gas supply lines (sample gas and zero
gas).
WARNING
SYSTEM SAFETY: This analyzer is typed as permanently connected
equipment and as such, must include a switch or circuit–breaker in the
installation. This switch must be supplied and installed by the customer (not
provided by Servomex) must be suitably located and easily reached, and must
be marked as the disconnecting device for the equipment.
3.4.5
Electrostatic Discharge
WARNING
Electrical Hazard, Personal Injury and Equipment Damage: The control of
ESD is critical when performing maintenance on the Servomex SpectraScan
Gas Analyzer. To control ESD, users should wear ESD wrist straps and other
grounding equipment between the analyzer and the skin. The wrist straps
should be tested for ground continuity often. Also, ground the analyzer
(through the frame chassis or the control rack) to protect against:
02400001A/0
Damage to sensitive electronic components, such as circuit boards
Electrostatic discharge
User injuries from electrical shock when contact is made with the
energized equipment.
19
SERVOTOUGH SpectraScan User Manual
3.4.6
Ergonomic Considerations
Ergonomic hazards are conditions that could result in harm to the operator or technician due to the
physical circumstances of operating or servicing the analyzer. Ergonomic hazards of the Servomex
SpectraScan Gas Analyzer include:
Lifting the analyzer
Stretching or maintaining an awkward position
WARNING
Personal Injury: The Servomex SpectraScan Gas Analyzer weighs
approximately 30 lbs (14Kg), presenting the possibility of personal injury if
care is not taken when lifting and moving this unit.
02400001A/0
20
SERVOTOUGH SpectraScan User Manual
4 UNPACKING THE ANALYZER
4.0 Component List
A standard Servomex SpectraScan Gas Analyzer is delivered with the following components:
02400001A/0
Servomex SpectraScan Gas Analyzer unit
Installation Manual
Calibration Verification Sheet
21
SERVOTOUGH SpectraScan User Manual
4.1 Removing the Analyzer from the Shipping Container
WARNING
Personal Injury: The Servomex SpectraScan Gas Analyzer weighs
approximately 30 lbs (14Kg), presenting the possibility of personal injury if
care is not taken when lifting and moving this unit.
NOTE
Do NOT discard any packaging material. All of the original packaging material
is required to return the unit to Servomex. The use of unsuitable packaging
material during shipping will void the analyzer’s warranty.
To remove the system from the shipping box, perform the following steps:
1. Open the top flaps on the cardboard shipping box and fold them out of the way.
2. Lift the unit together with the inner packaging materials, keeping the shock absorbing foam
pieces around the unit.
3. Set the unit (with the inner packaging still around it) on a flat workbench or table.
4. Carefully remove the foam packaging, the plastic bag and the desiccant (if applicable) from
the unit.
5. Return all of the packaging material to the box and store in a safe place for future use.
4.2 Inspection
Servomex has carefully packed the Servomex SpectraScan Gas Analyzer with the intent of
reaching its destination in perfect operating condition. Upon unpacking the unit, check for defects,
cracks, broken connectors, etc, to be certain that damage has not occurred during shipment.
If you find any damage, notify your freight carrier and Servomex immediately.
02400001A/0
22
SERVOTOUGH SpectraScan User Manual
5 MOUNTING CONSIDERATIONS
5.0 Environmental Requirements
The following considerations are described in this section:
Dimensions, Weight and Ventilation
Operating Temperature, Humidity and Pressure
Combustion Hazard
Power Requirements
Fittings Requirements
Figure 1 - The Servomex SpectraScan
Gas Analyzer
02400001A/0
23
SERVOTOUGH SpectraScan User Manual
5.0.1
Dimensions, Weight and Ventilation
The Servomex SpectraScan Gas Analyzer weighs approximately 30 lbs (14kg) and is designed to
be wall mounted. It requires a space 17 ¼” wide, 12” high (top to bottom) with an additional 6” of
clearance around the bottom for piping and electrical connections, and 7” wide (front to back) with
an additional 10 ½” clearance out the front to allow opening of the cover (when installing electrical
connections, or to make internal repairs or component replacements).
WARNING
Personal Injury: Improper installation can result in serious personal injury.
Make sure that the mounting structure can support a redundant weight factor
four times the total weight of the analyzer. If not, reinforce the structure before
installing the analyzer. The mounting structure must be made of metal, or at
minimum ½” plywood.
Dimensions of the mounting holes along the top and bottom edges of the analyzer are given in
Figure 2 below. The Servomex SpectraScan Gas Analyzer must be mounted with eight 3/8”-16 x 1”
UNC bolts and matching lock washers and nuts.
Figure 2 - Hole Dimensions of the Servomex SpectraScan Gas Analyzer
See Figure 3 - I/O panel of the Servomex SpectraScan Gas Analyzer on page 24 for a view of the
bottom of the analyzer, showing electrical and pipe connections.
02400001A/0
24
SERVOTOUGH SpectraScan User Manual
NOTE
The sample, power and signal connections require at least 6” of clearance for
safe operation.
WARNING
Personal Injury: The Servomex SpectraScan Gas Analyzer weighs
approximately 30 lbs (14Kg), presenting the possibility of personal injury if
care is not taken when lifting and moving this unit.
The Servomex SpectraScan Gas Analyzer should be mounted in a well-ventilated area.
WARNING
Personal Injury: Operate the Servomex SpectraScan Gas Analyzer in a wellventilated area to prevent the build-up of potentially harmful gases. Do NOT
vent sample cell exhaust into an area where it may be inhaled.
5.0.2
Operating Temperature, Humidity and Pressure
The Servomex SpectraScan Gas Analyzer can operate in a wide range of ambient temperature and
humidity, as shown in Table 14 - Environmental Specifications on page 72. The analyzer should be
operated in a non-condensing environment to avoid condensation on the optical windows in the gas
cell.
In addition to the ambient considerations, the Servomex SpectraScan Gas Analyzer has specific
requirements for the sample gas and zero gas being piped into it. Both sample gas and zero gas
have the same acceptable flow rate, pressure, humidity and temperature ranges. See Table 11 Sampling System Specifications on page 71. Refer to the Applications Manual for more sample
pressure recommendations.
There are overpressure considerations for the sample and zero gases entering the analyzer to
avoid injury and damage to the unit. See Section 7.0.3 Sampling Gas from a Pressurized Source on
page 51.
The sample and zero gas must be clean and free of particulates. A 10 micron particulate filter on
the supply side is recommended.
02400001A/0
25
SERVOTOUGH SpectraScan User Manual
5.0.3
Combustion Hazard
The Servomex SpectraScan Gas Analyzer is rated for certain combustible environments. Ensure
that the installation location of the analyzer aligns with the certifications listed on the product’s
labelling.
5.0.4
Power Requirements
The system is designed for 24 VDC operation. An optional AC Power Module can be installed at the
factory for use with 100-240 VAC. The main power board is fused for use with a 24 VDC supply.
Refer to Table 13 - Electrical Specifications on page 72 for more detailed information.
5.0.5
Fittings Requirements
The Servomex SpectraScan Gas Analyzer comes standard with ¼” male Swagelok tube fittings on
the sample inlet and exhaust ports. Only ¼” female Swagelok tube fittings should be used on the
incoming and outgoing sample lines, to mate with the Servomex SpectraScan sample inlet and
exhaust ports. Use of stainless steel tubing for the incoming and outgoing sample lines is
recommended. Instructions for installing the female Swagelok tube fittings onto the incoming and
outgoing sample lines are provided in Section 7 SAMPLE GAS CONNECTIONS on page 49.
Other fitting sizes and styles are available from Servomex to be installed on the inlet and exhaust
ports as a special order. The underlying connector is a ¼” NPT.
CAUTION
Equipment Damage: Provide adequate support for the tubing. The weight of the tubing must not
exert torque on the gas cell as it may cause the fittings to fail and leak.
All of the fittings requirements for the Servomex SpectraScan Gas Analyzer are summarized below.
Connector on the Analyzer
Required Fittings to
Mate with Analyzer
Additional Comments
Sample Inlet And Exhaust
Ports:
¼” male Swagelok
¼” female Swagelok
10 micron particulate filter on the
inlet side is recommended.
Electrical Conduit:
1” female
1” male conduit
Must maintain Ingress Protection
level of IP66, Type 4.
Purge Ports (use is optional):
¼” female NPTF
¼” male NPTF
Either side can be used as inlet.
Table 3 - Standard Fittings Used on the Servomex SpectraScan Gas Analyzer
5.1 Mounting Procedure
1. After choosing the appropriate space and structure for the Servomex SpectraScan Gas
Analyzer by satisfying all of the Environmental Requirements on page 23, prepare the unit
for mounting by clearing the mounting space, readying the analyzer (consider its weight and
02400001A/0
26
SERVOTOUGH SpectraScan User Manual
if you will need an assistant to support it while mounting), and having the mounting
hardware and appropriate tools nearby.
WARNING
Personal Injury: The Servomex SpectraScan Gas Analyzer weighs
approximately 30 lbs (14Kgs), presenting the possibility of personal injury if
care is not taken when lifting and moving this unit.
NOTE
The Servomex SpectraScan Gas Analyzer must be mounted with eight 3/8”16 x 1” UNC bolts and matching lock washers and nuts.
2. Consider the order of installation: mount the analyzer onto the mounting surface first; then
make electrical connections inside the box with the cover open.
3. Place the Servomex SpectraScan Gas Analyzer on the mounting structure and align the
eight mounting holes with holes in the structure. Support the analyzer if necessary.
4. Install eight 3/8”-16 x 1” UNC bolts through the mounting holes and through lock washers,
and secure with eight nuts. Tighten securely.
02400001A/0
27
SERVOTOUGH SpectraScan User Manual
This page intentionally left blank.
02400001A/0
28
SERVOTOUGH SpectraScan User Manual
6 ELECTRICAL CONNECTIONS
All of the required electrical connections are described in this section:
Electrical Power Source
Communication over Ethernet TCP/IP
NOTE
The Servomex SpectraScan Gas Analyzer must be mounted with eight 3/8”16 x 1” UNC bolts and matching lock washers and nuts.
WARNING
Personal Injury or Death: Wiring of the analyzer must be in accordance with
local regulations.
02400001A/0
29
SERVOTOUGH SpectraScan User Manual
6.0
I/O Panel Layout
Figure 3 - I/O panel of the Servomex SpectraScan Gas Analyzer
NOTE
The sample, power and signal connections require at least 6” of clearance for
safe operation.
The Red Circle in the figure above identifies the conduit fitting through which the power,
communication and external stream switching solenoid valve cables are to be connected.
The Green Circle in the figure above identifies where the sample inlet and exhaust lines are to be
connected.
The Blue Circles in the figure above identifies where the optional purge lines can be installed.
02400001A/0
30
SERVOTOUGH SpectraScan User Manual
Figure 4 - Analyzer with Cover Open
02400001A/0
31
SERVOTOUGH SpectraScan User Manual
6.0.1
Component Layout
Figure 5 - Interior of the Servomex SpectraScan Gas Analyzer
02400001A/0
NOTE
See Figure 6 for detail of SpectraScan internal components
32
SERVOTOUGH SpectraScan User Manual
Display
Cable
Enclosure Lid
Enclosure
Spare
Fuses
Ground Lug
Fuse
Plate
AC Power Module
(optional)
Circuit
Board
Earth
Ground
Lug
Motor Assembly
Power
Terminal
Block
Light Source
(beneath blue
cover)
Power
Cable
Sensor Head
(in blue cover &
TCP/IP partially beneath
Jack
plate and circuit
board)
Heater Wires
TCP/IP
Cable
Purge Port
Sample
Inlet
Conduit
Conduit
Sample
Exhaust Purge Port
Figure 6 - SpectraScan Interior Diagram
02400001A/0
33
SERVOTOUGH SpectraScan User Manual
6.1 Electrical Power Requirements
The Servomex SpectraScan Gas Analyzer operates at 24 VDC or 100-240 VAC depending on the
model configuration. Be sure to adhere to the Power Rating Label on the analyzer to determine
which power source to use, and connect the correct power source.
Figure 7 - Example Rating Label
WARNING
SYSTEM SAFETY: This analyzer is typed as permanently connected
equipment and as such, must include a switch or circuit–breaker in the
installation. This switch must be supplied and installed by the customer (not
provided by Servomex) must be suitably located and easily reached, and must
be marked as the disconnecting device for the equipment.
WARNING
Electrical Shock and Equipment Damage: The main circuit board in the
analyzer must have the proper fuse installed prior to connecting power.
Installing the wrong fuse in main circuit board may result in electric shock or
death and may result in damage to the analyzer. To avoid injury and
equipment damage, ensure the proper fuse is used. See Table 13 - Electrical
Specifications on page 72. Servomex provides the proper fuse and two spares
with every analyzer.
See Figure 6 - SpectraScan Interior Diagram on page 33.
02400001A/0
34
SERVOTOUGH SpectraScan User Manual
6.2 Wiring
All cables entering the analyzer (24 VDC or 120/240 VAC power and communication) must be
separately jacketed as they enter the analyzer to prevent miss-wiring. The power cabling used,
whether for 24 VDC operation, or for 120-240 VAC operation, must use a 3 conductor cable. The
ground wire in this 3 conductor cable must be electrically connected to the earth ground lug in the
analyzer.
Connection
Wire type
Power (AC)
600V, 3 wire, 16AWG, Jacketed
Insulation must be rated to at least 75°C
Power (DC)
600V, 3 wire, 16AWG, Jacketed
Ethernet (TCP/IP)
CAT 5 Standard 300V
Solenoids (DC)
600V, 3 wire, 18AWG
RS485 Data
Data Cable 24 AWG, 300V
Optional Heater
600V, 3 wire, 18AWG
Table 4 - Wire Types
02400001A/0
35
SERVOTOUGH SpectraScan User Manual
6.2.1
Power Wiring: 24 VDC
Figure 8 - Wiring the SpectraScan with 24 VDC
02400001A/0
36
SERVOTOUGH SpectraScan User Manual
To connect 24 VDC supply power to the analyzer, perform the following steps:
(Refer to Figure 8)
1. Open the cover of the SpectraScan using a large, flat-headed screwdriver turn the latch ¼ turn
counter-clockwise, and open the cover.
2. Bring a 3 conductor, separately jacketed cable with 24V supply into the enclosure through the
conduit fitting. Make sure that the power is not on (wires are not live) at this point.
WARNING
Maintain Ingress Protection: The Servomex SpectraScan unit has been
tested to an Ingress Protection level of IP66, Type 4. In order to maintain this
rating, a suitable 1” conduit connector must be used that will not compromise
the existing IP Rating.
3. Strip the cable jacketing back and secure the jacketed portion of the cable to the anchor point
provided inside the enclosure using a cable tie. Be sure to leave enough slack in the cable so
the red and black wires reach the power terminal block on the main board and the green wire
reaches the ground lug on the enclosure wall.
4. Strip 10mm (0.4”) of insulation from the positive (red) and negative (black) leads of the cable,
and also from the ground (green) wire.
5. Connect the positive (red) cable to the positive side of the terminal block on the main PCB and
the negative (black) cable to the negative side of the input power terminal block.
6. Connect Earth Ground:
a. Connect the green wire to the earth ground lug on the enclosure wall.
b. Check the two factory installed ground wires, at both ends, for loose connections;
resolve any that are found. Ensure that both are secured to the enclosure anchor
point with a cable tie. The factory installed ground wires are:
i. PCB to earth ground
ii. Enclosure lid to earth ground
7. Close and latch the cover.
02400001A/0
37
SERVOTOUGH SpectraScan User Manual
6.2.2
Power Wiring: 120-240 VAC
Figure 9 - Wiring the SpectraScan with 120-240Vac
02400001A/0
38
SERVOTOUGH SpectraScan User Manual
To connect 100-240 Vac power to the analyzer, perform the following steps:
(Refer to Figure 9)
1. Open the cover of the SpectraScan using a large, flat-headed screwdriver turn the latch ¼ turn
counter-clockwise, and open the cover.
2. Bring a 3 conductor, separately jacketed cable with AC supply into the enclosure through the
conduit fitting. Make sure the wires are not live at this point.
WARNING
Maintain Ingress Protection: The Servomex SpectraScan unit has been
tested to an Ingress Protection level of IP66, Type 4. In order to maintain this
rating, a suitable 1” conduit connector must be used that will not compromise
the existing IP Rating.
3. Strip the cable jacketing back and secure the jacketed portion of the cable to the anchor point
provided inside the enclosure using a cable tie. Be sure to leave enough slack in the cable so
the ends reach the input terminals on the AC Power Module.
4. Strip 10mm (0.4”) of insulation from the line (black), neutral (white) and ground (green) leads of
the cable.
5. Connect Earth Ground:
a. The terminal tabs on the AC Power Module have three positions:
Down: terminal locked closed
Middle: terminal closed with spring tension
Up (or to the left): terminal locked open
b. Connect the ground (green) wire to the Earth Ground terminal on the AC Power
Module. Do not disturb the factory installed power-supply-to-earth-ground wire that
is also inserted into the Earth Ground terminal on the AC Power Module.
c.
Check all three factory installed ground wires, at both ends, for loose connections;
resolve any that are found. Ensure two of the wires are secured to the enclosure
anchor point with a cable tie. The three factory installed ground wires are:
i. Power supply to earth ground
ii. PCB to earth ground (secured at anchor point)
iii. Enclosure lid to earth ground (secured at anchor point)
6. Connect the neutral (white) wire to the Neutral terminal on the AC Power Module.
7. Connect the line (black) wire to the Line terminal on the AC Power Module.
8. Close and latch the cover.
02400001A/0
39
SERVOTOUGH SpectraScan User Manual
6.3 Communication Connection
Figure 10 - Connecting the Ethernet TCP/IP communication cable
02400001A/0
40
SERVOTOUGH SpectraScan User Manual
To connect the communication cable to the analyzer, perform the following steps:
(Refer to Figure 10)
WARNING
Electrical Hazard, Personal Injury and Equipment Damage: Do NOT
perform any internal service on an energized or gas supplied analyzer. When
performing maintenance on the unit, it is necessary to disconnect the power
source from the unit, and turn off the gas supply lines (sample gas and zero
gas).
1. Turn off the main power, disconnect and turn off all gas supply lines that are connected.
2. Open the cover of the SpectraScan using a large, flat-headed screwdriver turn the latch ¼ turn
counter-clockwise, and open the cover.
3. Bring the communication cable into the enclosure through the conduit fitting.
WARNING
Maintain Ingress Protection: The Servomex SpectraScan unit has been
tested to an Ingress Protection level of IP66, Type 4. In order to maintain this
rating, a suitable 1” conduit connector must be used that will not compromise
the existing IP Rating.
4. Plug the communication cable into the network jack on the front edge of the PCB, as shown.
5. Secure the cable to the anchor points provided inside the enclosure using cable ties.
6. Close and latch the cover.
7. Restore power and gas supply, as applicable.
02400001A/0
41
SERVOTOUGH SpectraScan User Manual
6.4 Spectrascan Networking and Software Configuration
6.4.1
Ethernet and TCP/IP
Getting an IP address using DHCP
As delivered, the instruments are configured as DHCP clients. This means each instrument's IP
addresses given to it by an external DHCP server, typically a router running on your network. The
IP address is assigned dynamically when the instrument is powered up. To enable networking and
get a dynamic IP address, do the following:
1.
Attach the instrument to your network via Ethernet cable to the connector labelled
“ETHERNET” on the instrument's system board.
2.
Make sure your network has a DHCP server running.
3.
When the instrument is powered up, it will issue a request for an IP address via DHCP.
The DHCP server will answer the request and provide an IP address to the instrument. The
instrument will remember that IP address for as long as it remains powered up.
Note that the IP address assigned via DHCP will be lost if the instrument is powered down. The
next time you power up the instrument it may get the same IP address, or it may get a different one
depending upon the exact details of your network's configuration.
6.4.2
Finding the IP address of your instrument
Each instrument is programmed with a hostname. The hostname is of the form “oxsXXXXXX”,
where XXXXXX is the six-digit serial number assigned to the instrument. The serial number is
printed on a label attached to the bottom mount-plate in the vicinity of the stepper motor (Figure 11 Location of the serial number tag). The serial number is identified by the prefix S/N.
Figure 11 - Location of the serial number tag
02400001A/0
42
SERVOTOUGH SpectraScan User Manual
NOTE
The location of serial number tag is circled in red. If you look carefully, you
can see this instrument's serial number is 100072. Therefore, this
instrument's hostname is “oxs100072”.
Once you know the hostname of your instrument, you may find its IP address in the following way:
1.
On a Windows PC attached to the same sub-network as the instrument, open an MS DOS
command prompt.
2
In the command prompt, type “nslookup ”, where is the
instrument's hostname as described above.
3.
Nslookup will return all information about the instrument pertinent to networking, including
the IP address it has been assigned.
6.4.3
Setting a static IP address
Some customers prefer to assign a static IP address to their instruments. Assigning a static IP
address requires that you have physical access to the system board.
Connecting to the system board via RS-232 requires the following items:
Windows PC
A special serial cable from Servomex. On one end this cable is terminated by a connector
mating to the RS-232 connector on the system board. On the other end, the cable is
terminated by a DB-9 connector.
If your computer does not have an RS-232 port, you will need an RS-232 to USB converter.
The RS-232 side of the converter should be terminated with a DB-9 connector.
A terminal program running on your PC, such as Hyperterm or PuTTY.
You may configure your instrument to use a static IP address in the following way:
1. Open the instrument, and locate the MPU debug connector. The connector is shown in
Illustration 2. This connector provides access to the MPU via an internal RS-232 port.
02400001A/0
43
SERVOTOUGH SpectraScan User Manual
Figure 12 - Location of debug RS-232 connector on MPU
2.
Connect a Windows PC to the MPU debug connector using the SpectraScan RS-232 cable
(and an RS-232 to USB converter, if required).
3.
On your PC, determine which COM port is associated with the serial connection. Do this by
opening the Windows device manager and looking at the COM port configuration.
4.
On your PC, run your terminal program and configure it to use the COM port you found in the
device manager. Also configure the terminal program with the following settings:
Baud rate = 115200
Data bits = 8
Stop bits = 1
Parity = None
Flow control = None. (No software nor hardware flow control.)
5.
In the terminal program, start a terminal session to the MPU. Hit a few times and you
will get a command prompt from the MPU. The command prompt is “#”.
6.
At the command prompt, type the command “ip-config” followed by . When the
program starts it will tell you the current IP configuration of the unit. Also, if the unit is already
configured for static IP address, it will give you the option to leave the program.
7.
Next, the program will give you three options to modify the IP address settings. Choose option
2, “static IP”.
8.
The program will
9.
When prompted, type in your desired IP address is DD.DD.DD.DD format. (Here, DD signifies
the digits comprising each of the four decimal numbers making up the IP address.)
10. Next, you will be prompted for the desired netmask. Enter it using the DD.DD.DD.DD format.
02400001A/0
44
SERVOTOUGH SpectraScan User Manual
11
.Next, you will be prompted for the desired gateway address. Enter it using the DD.DD.DD.DD
format.
12. Next, you will be prompted for the desired broadcast address. Enter it using the
DD.DD.DD.DD format.
13
.Finally, you will be asked to reboot and accept the changes, or to cancel. If you have entered
all information correctly, choose reboot. The system will then store the new IP configuration
data, shut down, and then restart.
14. The system requires roughly 1 minute to reboot. Once the system has rebooted, you will be
presented with a command prompt, “#”.
15. You may now disconnect the RS-232 cable from the MPU debug port and close the instrument.
A screenshot showing a typical session is shown in Illustration 3 below. In this session, an
instrument is configured to use the static IP address 192.168.0.29 with netmask 255.255.255.0,
gateway 192.168.0.1 and broadcast IP address 192.168.0.255.
02400001A/0
45
SERVOTOUGH SpectraScan User Manual
Figure 13 - Sample session in which a static IP address is set using PuTTY to
communicate with the MPU.
6.4.4
RS-422 connection
The RS-422 link uses a 4-wire connection, 2 wires for receive and 2 wires for transmit. The RS-422
port supports only a point-to-point link. Do not bridge the receive and transmit pairs – the
instrument cannot be used in multi-drop bus configurations.
Connecting the instrument's RS-422 port to an external RS-422 device requires a 5 conductor
cable. The cable should include two twisted pairs for the receive and transmit data, and a single
wire for data ground. The configuration is shown in Drawing 1. Connection to the system board is
made using the connector labelled “RS-485” at the corner of the board.
02400001A/0
46
SERVOTOUGH SpectraScan User Manual
Figure 14 - Identification of RS-422 signals in instrument
This is shown in Illustration 4. Connect the cable to the individual push-down connectors as shown
in the figure. (NOTE: Although the board label says “RS-485”, this port will not function in multidrop bus mode. It will only work as a point-to-point link with a 4-wire RS-422 device.)
Figure 15 - Location of RS-422 connector on system
board
The RS-422 port parameters are as follows:
Baud rate = 57600
Data bits = 8
Stop bits = 1
Parity = None
02400001A/0
47
SERVOTOUGH SpectraScan User Manual
6.4.5
Modbus register
The SpectraScan communicates to the external controller via Modbus. The instrument acts as a
Modbus server, the external controller is the client. Modbus runs either over RS-422 or Ethernet
interfaces. The instrument is configured in the following way:
Modbus protocol: RTU
Modbus device address = 1
The Modbus register map used by the SpectraScan is shown below.
Table 5 - SpectraScan Modbus register
02400001A/0
48
SERVOTOUGH SpectraScan User Manual
7 SAMPLE GAS CONNECTIONS
SpectraScan Gas Analyzer
10 Micron
Particulate
Filter
Purge Gas
Supply
Manual
Valve
Zero Gas
Supply
Manual
Valve
Purge Gas
Exhaust
Sample Gas
Exhaust
Pressure Relief
Mechanism
NC
Zero
Gas
Solenoid
NO
Sample Gas
Solenoid
Manual Valve
Sample Gas
Supply
Figure 16 – Typical Gas Connection Configuration
02400001A/0
NOTE
The specific configuration of sample, purge and zero gases may vary
depending on application.
49
SERVOTOUGH SpectraScan User Manual
7.0 Sample Line Connections
7.0.1
Sampling Ports
Both sample intake and exhaust ports are fitted with male ¼” Swagelok® compression tube fittings.
Only properly mated, female ¼” Swagelok® compression fittings should be used on the sample
intake and exhaust lines.
CAUTION
Equipment Damage: Restricting the gas flow on either the intake or the exhaust side may cause
the gas cell to over pressurize, causing damage to the analyzer. Care should be taken to remove
any port caps or other flow barriers before connecting any sample line or switching on the
analyzer. Also make sure the exhaust routing is appropriate for your specific application.
CAUTION
Equipment Damage: Provide adequate support for the tubing. The weight of the tubing must not
exert torque on the gas cell as it may cause the fittings to fail and leak.
7.0.2
Preparing Sample Gas Lines
Instructions for installing female ¼” Swagelok tube fittings on the sample incoming and exhaust
lines:
1. Insert the tubing into the ¼” female Swagelok tube fitting. Tubing should rest firmly on the
shoulder of the fitting. The nut should be finger tight.
2. Hold the fitting body with a backup wrench and tighten the nut 1¼ turns.
02400001A/0
50
SERVOTOUGH SpectraScan User Manual
7.0.3
Sampling Gas from a Pressurized Source
When the gas is to be sampled from a pressurized source, such as a pressurized line, a pressure
relief mechanism should be employed so as not to force or restrict the flow of gas into the unit.
Refer to the Applications Manual for sample pressure recommendations. See Table 11 - Sampling
System Specifications on page 71 for allowable pressure, temperature and flow rate.
WARNING
Personal Injury and Equipment Damage: Sample allowable pressure must
not be exceeded, or may cause the gas cell to over pressurize. Care should
be taken to review Table 11 - Sampling System Specifications on page 71.
CAUTION
Equipment Damage: Restricting the gas flow on either the intake or the exhaust side may cause
the gas cell to over pressurize. Care should be taken to remove any port caps or other flow
barrier before switching on the analyzer. Also make sure the exhaust routing is appropriate for
your specific application.
CAUTION
Equipment Damage: Pressure relief mechanism should be employed when sampling gas from
a source other than the ambient in order to avoid damage to the analyzer.
7.0.4
Cleanliness of Sample Stream
CAUTION
Equipment Damage: The sample stream must be clean of particles that could block the light
path and potentially accumulate on the windows, and potentially damage the unit. It is ultimately
the user’s responsibility to ensure a clean sample stream. We recommend an inline 10 micron
particulate filter for most applications. Contact Servomex for more information.
02400001A/0
51
SERVOTOUGH SpectraScan User Manual
7.0.5
Stream Switching Using External Solenoid Valves
The Servomex SpectraScan Gas Analyzer is equipped with three relays connected to the 24 VDC
supply, which may be used to control external solenoid valves for the purpose of stream switching.
By default, the first two relays are used for automatic zeroing: upon initiating, a software “zero”
command the relays energize, providing 24 VDC power to external solenoid valves. Upon
completion of the zeroing process, the relays return to their default “0” state. The analyzer supplies
power to the solenoid valves at the following specifications:
Voltage: 24 VDC
Maximum current: 0.5 A
WARNING
Personal Injury: Solenoids that will be driven from the Servomex
SpectraScan solenoid output must be safety agency approved.
02400001A/0
52
SERVOTOUGH SpectraScan User Manual
The diagram below depicts the recommended pipe work and electrical connections to utilize the
stream switching relays. Note that solenoid labels “NC” and “NO” denote normally closed and
normally open respectively. Solenoid lines 1, 2, 3 and 4 should be connected to the rightmost
connector bank on the front-edge of the PCB, to connector ports SOL1, GND, SOL2, and GND
respectively, as shown in the diagram below. Cables should be anchored with cable ties inside the
enclosure and exit the enclosure through the electrical conduit (along with the power and
communication cables).
Figure 17 - Recommended Connections for Inlet Stream Switching
02400001A/0
53
SERVOTOUGH SpectraScan User Manual
7.0.6
Procedure for Connecting the Sample Lines
Refer to Figure 16
WARNING
Electrical Hazard, Personal Injury and Equipment Damage: Do NOT
perform any internal service on an energized or gas supplied analyzer. When
performing maintenance on the unit, it is necessary to disconnect the power
source from the unit, and turn off the gas supply lines (sample gas and zero
gas).
1. Turn off the main power disconnect and turn off all gas supply lines.
2. Remove the port caps and any other obstructions from the sample path (inlet and exhaust
ports).
3. Install the appropriate fittings on the sample supply and exhaust lines (see Preparing
Sample Gas Lines on page 50).
4. Install a pressure relief mechanism if necessary (if sampling from a pressure other than
ambient).
5. Ensure a clean sample stream – install a 10 micron particulate filter on the inlet side of the
sample plumbing if necessary.
6. Install/connect stream switching lines and solenoids if applicable.
7. Connect sample supply and exhaust lines to the Servomex SpectraScan sample inlet and
exhaust ports.
8. Ensure proper sample gas and zero gas temperature and conditioning. In general, sample
conditioning is application dependent, see the Application data for more specific
information. However, samples entering the Servomex SpectraScan Gas Analyzer must
be “clean and dry” free of liquids and particulates. Sample cell is heated to 60°C. Any
sample above 50°C may have condensation or overheating risks.
CAUTION
Equipment Damage: The sample stream must be clean of particles that could block the
light path and potentially accumulate on the windows, and potentially damage the unit. It is
ultimately the user’s responsibility to ensure a clean sample stream. We recommend an
inline 10 micron particulate filter for most applications. Contact Servomex for more
information.
9. Check sample path (lines and connections) and zero gas path for leaks, by alternately
turning one on and the other off. Repair or replace to resolve any leaks that are found; the
Servomex SpectraScan Gas Analyzer must be operated in a leak-free environment.
10. Turn on the main power (at the disconnect).
11. Set supply and exhaust valves in the proper configuration for remote operation.
02400001A/0
54
SERVOTOUGH SpectraScan User Manual
7.1 Enclosure Purge Connections
7.1.1
Purging of Enclosure Compartment
Certain installations may require the use of purge gas to flow through the enclosure. There are two
female ¼” NPTF ports located on the analyzer for this purpose. (See Figure 3). To add a purge to
the enclosure, remove the plugs installed at the factory on the purge inlet and purge outlet ports
(either port can be used for inlet or outlet). Connect purge gas lines to the ports using male ¼”
NPTF connectors, or use fittings to adapt the purge ports on the analyzer to the connectors on the
gas purge lines. One purge port on the analyzer (acting as an inlet) should be connected to a
continuous supply of clean and dry inert gas such as nitrogen or air. The other port should be
connected to an exhaust line for the purge gas. Make sure the exhaust line is open, open the purge
gas line, and check for leaks.
WARNING
Personal Injury or Death: Do not connect sample lines to the purge fittings
as you could fill the enclosure and local area with potentially harmful or
explosive gases, exposing the user and others in the vicinity to the risk of
injury or death.
.
02400001A/0
55
SERVOTOUGH SpectraScan User Manual
This page intentionally left blank.
02400001A/0
56
SERVOTOUGH SpectraScan User Manual
8 INSTALLATION CHECKLIST
The following list gives a brief summary of all the steps needed to correctly install the Servomex
SpectraScan Gas Analyzer. It is not a substitution of the procedures, precautions and hazard
definitions that appear elsewhere in this document.
NOTE
Read this manual in its entirety before attempting to perform any operations
on the Servomex SpectraScan system.
1. Reviewed System Safety beginning on page 12.
2. Unpacked Servomex SpectraScan Gas Analyzer from shipping crate and inspected for
damage.
3. Chose appropriate mounting space and structure considering all Environmental
Requirements beginning on page 23.
4. Mounted the analyzer securely on the mounting structure.
5. Connected power source and communication cabling.
6. Connected sample inlet and exhaust lines, zero gas supply line and stream switching circuitry
as applicable.
7. Tested all gas paths for leaks and resolved.
8. Connected enclosure purge system (optional) and tested for leaks.
9. Supply gases properly conditioned for temperature, pressure, cleanliness and flow rate.
10. Supply and exhaust valves and electrical power disconnect are set in the proper
configuration for remote operation.
02400001A/0
57
SERVOTOUGH SpectraScan User Manual
This page intentionally left blank.
02400001A/0
58
SERVOTOUGH SpectraScan User Manual
9 VERIFICATION
The purpose of this procedure is to verify span accuracy upon installation of the Servomex
SpectraScan Gas Analyzer. This procedure is also performed at periodic intervals throughout the
effective life of the unit. This procedure has three parts: Setup, Zeroing and Span Verification.
Zeroing is a measurement on a zero (inert) gas. Span verification is a measurement on the
customer’s trusted, certified mixture(s).
9.0 Procedure
9.0.1
Setup
1. Communicate with the analyzer via Modbus/TCP: Refer to the Modbus register map
in the Applications Manual to be used to configure customer’s Modbus Master
Program (not provided by Servomex.)
2. Ensure that the Servomex SpectraScan Gas Analyzer is sufficiently warmed-up.
Warm-up: The unit’s interior is temperature-controlled and will only meet measurement accuracy
specifications when sufficient time is allowed for the unit to meet and stabilize to the designed
operating temperature. The red “Warm-Up” light on the display will remain on until the unit is
warmed-up. Warm-up time for the unit is approximately 30 minutes to 2 hours depending on the
ambient temperature. Do not begin to take data until the “Warm-up” light goes off and the “Run”
light (green) comes on. LED codes are as follows:
LED Label on
System Cover
Functional Description
Power
Lights green when power is connected.
Warm-up
Lights red while warming up, goes out when operational
temperature is reached.
Run
Lights green when operational temperature is reached.
Fault
Lights red when a fault is detected in the system.
Table 6 - Diagnostic LED Indications
.
02400001A/0
59
SERVOTOUGH SpectraScan User Manual
9.0.2
Zeroing
See
Stream Switching Using External Solenoid Valves on page 52 for information on controlling the
external valves.
1. Purge the cell with the appropriate zero gas: turn off the sample gas and turn on the zero
gas (through software). Refer to the Application data for the appropriate Zero gas
specifications.
2. Make sure that the gas cell is completely purged with the Zero gas by allowing an extra
amount of time in purging (See Table 12 - Response Time Versus Gas Flow Rate).
3. Initiate the command “Zero” from Modbus Master or controller.
(On the Test HMI software, a “Zero” toggle is on the “Configuration” tab.)
9.0.3
Span Verification
Purge the cell with your trusted, certified mixture of sample gas, or verification mixture as follows:
1. Connect the verification gas of choice to the sample inlet supply line if sourcing verification
gas from a cylinder.
2. Turn off the zero gas and turn on the sample gas (either through software in the case of
automated sample-switching configuration or manually).
3. Start collecting data.
4. Wait until data values from the analyzer stabilize. See Table 12 - Response Time Versus
Gas Flow Rate for more information.
5. Record values. Compare data with known values of your trusted, certified verification
mixture.
6. Complete Verification procedure or adjust span values as needed (see next section below).
9.1 Re-spanning
The customer may modify the factory set span values (default values of “1”) for any measurement
channel. This span modification can be done via Modbus/TCP. The span values are multiplied with
the original measured values to produce the final measurement values. For example, adjusting the
span value of a certain channel to 1.05 would adjust the measurement value up by 5%. Similarly,
adjusting the span value to 0.95 would adjust the measurement value down by 5%. Note that the
new span values are NOT stored in the analyzer’s memory, rather, it relies on the Modbus master
program (customer control system) to supply the information.
02400001A/0
60
SERVOTOUGH SpectraScan User Manual
10 MAINTENANCE
The Servomex SpectraScan Gas Analyzer is designed to be maintenance free for extended periods
of time. In a normal installation, it will operate unattended in its intended environment for 18 months
with no physical intervention. Typically every 18 months, the spectrometer light source needs to be
replaced.
Recommended maintenance and operation verification of the Servomex SpectraScan Gas Analyzer
consists of:
Monthly re-zeroing (performed remotely through software).
Light source assembly replacement (every 18 months or when light indicator shows < 40%
output). The replacement procedure is described in Section 10.0.
Gas cell window replacement if necessary when prompted by fault signal.
02400001A/0
61
SERVOTOUGH SpectraScan User Manual
10.0 Light Source Assembly Replacement Procedure
Light source replacement is recommended every 18 months or when the light indicator shows <
40% output. Neither the light source nor its circuit board can be repaired.
The light source assembly is a small circuit board with a light source under it, which attaches to the
top of the sensor head via a mounting hole for the light source and four bolts with lock washers.
Access only requires peeling back a corner of the blue bag surrounding the cell, and Velcro makes
this easy. Removal of the assembly is executed by disconnecting its connector and removing the
four bolts that hold it into place. Heat sink compound is applied to the light source prior to inserting
it into the cell.
WARNING
Electrical Hazard, Personal Injury and Equipment Damage: Do not perform
maintenance on an energized or gas supplied analyzer. When performing any
procedure that involves opening the unit, it is necessary to disconnect the
power source from the unit, and turn off the gas supply lines (sample gas and
zero gas).
1. Turn off all gas supply lines and turn off power to the analyzer before performing any
service procedure.
2. Open the cover of the SpectraScan using a large, flat-headed screwdriver turn the latch ¼
turn counter-clockwise, and open the cover.
3. Expose the light source assembly from underneath the blue bag:
Refer to Figure 5 - Interior of the Servomex SpectraScan Gas Analyzer on page 33 to
identify the location of the light source assembly.
4. Peel back the cover by opening the Velcro on the upper and right edges of the blue bag, as
shown in Figure 18.
02400001A/0
62
SERVOTOUGH SpectraScan User Manual
Figure 18 - Accessing the Light Source
5. Remove the connector with the green wires from the bottom edge of the light source circuit
board as shown in Figure 19. The black connector has a squeeze tab to prevent accidental
release of the connector.
Figure 19 - The Light Source Circuit Board
6. Using a 3/32” hex wrench, loosen and remove the four bolts and four lock washers at the
corners of the light source circuit board. Lift the blue light source circuit board up and out of
its mounting hole. This circuit board with the light source on the other side of it is the light
source assembly (see Figure 20).
7. The light source assembly shall be replaced as a whole. It shall not be reduced to smaller
components or repaired. The old assembly can be discarded in accordance with
appropriate local regulations.
02400001A/0
63
SERVOTOUGH SpectraScan User Manual
Figure 20 - The Light Source Assembly
8. Apply heat sink around the perimeter of the new light source, and push the assembly back
into the mounting hole provided on the sensor head, making sure to orient it properly
(connector end toward I/O panel).
Figure 21 - Mounting Hole for Light Source
9. Replace the four bolts and four lock washers to secure the circuit board to the sensor head.
Tighten the bolts in a cross pattern.
10. Replace the connector back into its socket on the light source circuit board. The tab will
snap in place to lock the connector into position.
11. Re-close the blue bag by sealing the Velcro strip along its edges.
12. Close the cover and lock by rotating the latch.
02400001A/0
64
SERVOTOUGH SpectraScan User Manual
13. Turn on power and gas supply.
14. Wait for at least 30 minutes and until the cell temperature stabilizes at the set operating
temperature (indicated by the green “Run” light on the display).
15. Perform Verification (refer to Section 9 Verification).
02400001A/0
65
SERVOTOUGH SpectraScan User Manual
This page intentionally left blank.
02400001A/0
66
SERVOTOUGH SpectraScan User Manual
11 TROUBLESHOOTING
See Table 6 - Diagnostic LED Indications for a general description of each indicator status. Below
are some common problems/symptoms and possible causes.
Problem
Possible cause(s)
Power LED not lit
No or insufficient power. Check Fuse
Warm-up light still lit after 2 hours of warm up
period
Environment is too cold to allow stabilization at
the operating temperature
Negative concentration reading
Zeroing on sample gases or insufficient purging
prior to zeroing
Excessive drift
Analyzer has not fully warmed up or sensor
head temperature is higher than its intended
operating temperature (default 60° C)
Unable to communicate with the analyzer
Incorrect IP address
Low light indicator (<50%)
Dirty windows or light source has degraded
No measurement response or unchanging
measurement values
Unconnected or improperly connected sample
line, blocked flow
Slow measurement response
Low flow rate due to blockage
Table 7 - Troubleshooting
02400001A/0
67
SERVOTOUGH SpectraScan User Manual
This page intentionally left blank.
02400001A/0
68
SERVOTOUGH SpectraScan User Manual
12 APPENDIX A: PRODUCT SPECIFICATIONS
12.0 Measurement Specifications
Measurement
Technique
Light absorption spectroscopy
Gas Cell Volume / Path
Length
100 mL / 35 cm
Zero Drift
<0.2% of full scale (FS) per month (typical)
Intrinsic Error
(Accuracy)
<1% FS (typical).
Higher accuracy available upon request
Repeatability
<0.1% (typical)
Table 8 – Measurement Specifications
02400001A/0
69
SERVOTOUGH SpectraScan User Manual
12.1 Physical Specifications
Figure 22 - Dimensions of SpectraScan
Size
17.25”x 12” x 7” deep
Weight
28 lbs. (14 kg)
Installation
Wall Mounted
See Dimensions, Weight and Ventilation on page 24 for installation
considerations
Vibration
Immunity
Resistant to low amplitude vibration
Table 9 – Physical Specifications
12.2 Communications Specification
Communication
Modbus TCP/IP Ethernet
Table 10 - Communications Specifications
02400001A/0
70
SERVOTOUGH SpectraScan User Manual
12.3 Sampling System Specifications
Item
Value
0 to 15 ltr/min (higher ranges available upon request)
Allowable Flow Rate
0 to 30 psig (standard)
Allowable Gas Cell Pressure
Allowable Sample
Temperature and Humidity
Material Of Wetted Parts
0 to 80 psig (optional)
Refer to product labelling for model-specific information.
0 to 50°C (32 to 122°F) standard
5 to 95% rH non-condensing
Contact Servomex for higher temperature requirement
Nickel plated or anodized aluminium gas cell, (stainless steel gas
cell is optional)
Stainless steel tubing and fittings, Viton O-rings,
BK7/quartz/sapphire/CaF2/ZnSe optical windows
Table 11 - Sampling System Specifications
FLOW RATE
T90 (SECONDS)
T95 (SECONDS)
T99 (SECONDS)
0.5 SLM
25
35
60
1 SLM
20
25
40
5 SLM
4
5
6
Table 12 - Response Time Versus Gas Flow Rate
02400001A/0
71
SERVOTOUGH SpectraScan User Manual
12.4 Electrical Specifications
Power Rating and Fuse Information
Base Input Power
24 Vdc, 5.0 A max
Optional Input Power
100-240 Vac (+/-) 10%, 50-60 Hz, 3.0 A max*
Fuse (24Vdc replaceable)
125V, 10A
Table 13 - Electrical Specifications
*See Table 4 - Wire Types on page 35 for important wire rating information.
12.5 Environmental Specifications
Ambient Operating Temperature/Humidity Range
-20 to 50°C (-4 to 122°F)
(5 – 95% rH non-condensing)
System Cooling
None required
Storage/Shipping Temperature/Humidity
Range
-20 to +60°C (-4 to 140°F)
(5 – 95% rH non-condensing)
Altitude
0 – 2000m (0 – 6560 ft)
Overvoltage Category
Category II
Pollution Degree
Pollution Degree 3
Dust Protection Rating (Except for Gas Sample
Path)
IP rating = 66 (unit interior is protected
against dust)
Table 14 - Environmental Specifications
12.6 Manual Revision History
Draft
Description
Effective Date
0
Initial Release
May-13
02400001A/0
72
SERVOTOUGH SpectraScan User Manual
13 INDEX
combustion hazard, 26
communication requirements, 70
connections
communication, 40
electrical, 29
plumbing, 41–55
power, 34
solenoid relays, 53
wire types, 35
dust protection rating, 72
Fittings Requirements, 26
fuse, 72
hazard notices, 7
hazards
over pressurizing, 18
ventilation, 24
weight, 24
LED codes, 59
measurement principle, 9
power
AC, 38
DC, 36
power rating, 72
power rating label, 34
power requirements, 26, 34, 52, 72
allowable gas cell, 71
purging, 55
regulation, 12
sample
cleanliness, 51
ports, 51
pressurized source, 51
solenoids, 52
specifications
communication, 70
electrical, 72
environmental, 72
mounting, 23–27
physical, 70
sampling, 71
wiring, 35
standards in the industry, 12
stream switching, 52
symbols found on the analyzer, 8
temperature
allowable sample range, 72
operating range, 72
shipping and storage range, 72
ventilation, 24, 25, 55
warm-up time, 15
pressure
02400001A/0
73
SERVOTOUGH SpectraScan User Manual
14 CONTACT DETAILS
Servomex Group Limited
Jarvis Brook
Crowborough
East Sussex
TN6 3FB
United Kingdom
TEL +44 (0)1892 652181
Fax +44 (0)1892 662253
www.servomex.com
02400001A/0
NOTE
All analyzers returned to Servomex or one of its appointed agents for
servicing, or any other purpose must be accompanied by a completed
decontamination certificate.
74