Servomex SERVOTOUGH SpectraExact (2500 series) Installation Manual
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
PROCESS ANALYSERS
SERVOTOUGH SpectraExact
(2500 series)
Process Gas Analysers
Installation Manual
Part Number:
Revision:
Language:
02500005D
0
UK English
This page intentionally blank
Servomex 2500 Series
Process Gas Analysers
Installation Manual
Ref:02500/005D/0
Order as part 02500005D
WARNING, CAUTIONS AND NOTES
This publication includes WARNINGS, CAUTIONS and NOTES which provide,
where appropriate, information relating to the following:
•
WARNINGS: Hazards which will result in personal injury or death.
•
CAUTIONS: Hazards which will result in equipment or property damage.
•
NOTES: Alert the user to pertinent facts and conditions.
WARNING
•
LETHAL VOLTAGES: THE ELECTRICAL POWER USED IN THIS
EQUIPMENT IS AT A VOLTAGE HIGH ENOUGH TO ENDANGER LIFE.
•
BEFORE CARRYING OUT MAINTENANCE OR REPAIR THE EQUIPMENT
MUST BE DISCONNECTED FROM THE ELECTRICAL SUPPLY. TESTS
MUST BE MADE TO ENSURE THAT DISCONNECTION IS COMPLETE.
•
IF FOR ANY REASON THE POWER SUPPLY CANNOT BE
DISCONNECTED, FUNCTIONAL TESTING, MAINTENANCE AND REPAIR
OF THE ELECTRICAL UNITS IS ONLY TO BE UNDERTAKEN AS A LAST
RESORT AND MUST BE CARRIED OUT BY PERSONS FULLY AWARE
OF THE DANGER INVOLVED.
NOTE
This instruction manual for the 2500 Series range of analysers covers installation,
routine maintenance and fault diagnosis procedures, and it should be thoroughly
read and retained by the personnel responsible.
i
ii
Table of Contents
SECTION 1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1
1.1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1
1.2
General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2
1.3
Basic Construction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.3
1.4
Options and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
1.4.1
Heated Sample Cell – Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
1.4.2
Heated Sample Cell – Steam . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.5
1.4.3
Optional Output Cards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.5
1.4.4
Compensation Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.5
1.4.5
Digital Communications Options . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.6
1.4.6
Application Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.6
SECTION 2 INSTALLATION – GENERAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1
2.1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1
2.2
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1
2.2.1
Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1
2.2.2
Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1
2.2.3
Unpacking and Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1
2.3
Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.2
2.4
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.3
2.4.1
Positioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.3
2.4.2
Mounting Details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.4
2.5
Instructions Specific to Hazardous Area Installations . . . . . . . . . . . . . . . . . . . 2.7
2.6
ATEX Label Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.8
SECTION 3 INSTALLATION – ELECTRICAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1
3.1
Electrical Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1
3.2
Signal Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6
3.2.1
Analogue Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.7
3.2.2
Analogue Output Link Selections on Sensor Interface
Board (SIB) PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.9
3.2.3
Relay Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.10
3.2.4
Relay Output Link Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.10
3.3
Control Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.11
3.3.1
Range Change Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.12
3.3.2
Autocalibration Initiation Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.12
3.3.3
Password Keyswitch Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.12
3.3.4
Sample Flow Sensor Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.13
3.3.5
Solenoid Valve Drives . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.13
3.3.6
Externally Powered Solenoid Valve Connections . . . . . . . . . . . . . 3.14
3.3.7
Internally Powered Solenoid Valve Connections . . . . . . . . . . . . . . 3.15
3.4
Digital Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.16
3.4.1
RS-232 Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.16
3.4.2
Modbus RS-485 Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.23
3.4.3
Modbus TCP Ethernet Connection . . . . . . . . . . . . . . . . . . . . . . . . 3.23
3.5
Pressure Transducer Connections – (If Supplied) . . . . . . . . . . . . . . . . . . . . . 3.24
iii
SECTION 4 INSTALLATION GAS CONNECTIONS . . . . . . . . . . . . . . . . . . . . . . . 4.1
4.1
Purge Connections (EU1/EX1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.1
4.1.1
General Purge Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.1
4.2
End Boss Purge Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.3
4.2.1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.3
4.2.2
End Boss Scrubbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.3
4.2.3
End Boss Vents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4
4.2.4
End Boss Purge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4
4.3
Steam Heated Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.6
4.4
Process Stream Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.6
4.4.1
Basic Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.6
4.4.2
High Integrity Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.6
4.5
Power-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.6
SECTION 5 FAULT DIAGNOSIS AND CELL MAINTENANCE . . . . . . . . . . . . . . 5.1
5.1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1
5.2
Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1
5.3
General Fault Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.4
5.4
Serious Fault (Shutdown) Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.6
5.5
Cell Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.6
5.6
Routine Leak Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.11
SECTION 6 SPARES LISTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1
6.1
Spares List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1
6.1.1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1
6.1.2
Sample Cell Spares . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.2
6.1.3
Source Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3
6.1.4
Detector Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3
6.1.5
Windows . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.4
6.1.6
Scrubbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.4
6.2
Recommended Spares . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.5
SECTION 7 INSTRUMENT SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1
7.1
Generic 2500 Series Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1
7.1.1
Environmental Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1
7.1.2
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1
7.1.3
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1
7.1.4
Performance Characteristics (each component) . . . . . . . . . . . . . . . 7.2
7.1.5
Performance Characteristics – EMC . . . . . . . . . . . . . . . . . . . . . . . . 7.2
7.1.6
Sample Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.3
7.1.7
Sample and Ambient Performance . . . . . . . . . . . . . . . . . . . . . . . . . 7.3
7.1.8
Solenoid Valve Relays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.3
7.1.9
Analogue Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
7.1.10
Serial Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
7.1.11
Modbus RS-485 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
7.1.12
Ethernet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.5
7.1.13
Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.5
7.1.14
Ingress Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.5
7.1.15
Optical Purge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.5
7.1.16
Alternative Purge Controller (2500 EU1 & EX1 Only) . . . . . . . . . . . 7.5
iv
SECTION 8 CE MARKING AND OTHER SAFETY APPROVALS . . . . . . . . . . . . 8.1
8.1
EMC and Electrical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.1
8.2
ATEX Directive and other non-European Hazardous Area approvals . . . . . . . 8.2
APPENDIX A Modbus Data Mapping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A.1
A.1
Floating Point Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A.1
A.2
Measurement Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A.2
A.3
Measurements Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A.2
A.4
Measurement Diagnostics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A.3
A.5
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A.3
A.6
Fault Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A.4
A.7
Autocalibration Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A.5
A.8
Autocalibration Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A.6
A.9
Autocalibration Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A.7
APPENDIX B Equipment Protection Levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B.1
APPENDIX C Disposal In Accordance With the WEEE Directive . . . . . . . . . . . C.1
v
List of Figures
Figure 1.1
Figure 1.2
Figure 2.1
Figure 2.2
Figure 3.1a
Figure 3.1b
Figure 3.2
Figure 3.3
Figure 3.4
Figure 3.5
Figure 3.6
Figure 4.1
Figure 4.2
Figure 5.1
Figure 5.2
2500 Overall View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2
2500 Construction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
2500 Positioning Restrictions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.5
2500 Mounting Details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.6
2500 AC Power and Valve Connections (Purged Filter) . . . . . . . . . . . 3.3
2500 AC Power and Valve Connections (Non-Purged Filter) . . . . . . . 3.4
2500 Series Terminal Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.5
Signal Cable Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6
SIB and Optional Output PCB's . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.7
Autocalibration Valve Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . 3.16
Typical Pressure Transducer Assembly . . . . . . . . . . . . . . . . . . . . . . . 3.25
Purge Connection Schematic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.1
Cell End Boss Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.5
General Cell Construction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.7
Suppressor Mounting Details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.8
List of Tables
Table 3.1
Table 3.2
Table 3.3
Table 3.4
Table 3.5
Table 3.6
Table 3.7
Table 3.8
Table 3.9
Table 3.10
Table 5.1
Table 5.2
Table 5.3
Table 5.4
Table 6.1
Table 6.2
Table 6.3
Table 6.4
Table 6.5
Table 6.6
Table 6.7
Table 7.1
Table 7.2
Table 7.3
Table 7.4
Table 7.5
Table 7.6
Table 8.1
vi
Analogue Output Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.8
Relay Output Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.10
External Control Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.11
Externally Powered Solenoid Valve Connections . . . . . . . . . . . . . . 3.14
Internally Powered Solenoid Valve Connections . . . . . . . . . . . . . . 3.15
Serial Data Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.17
Fault Number Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.21
ASCII Characters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.22
RS232 Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.22
Pressure Transducer Connections . . . . . . . . . . . . . . . . . . . . . . . . . 3.23
Diagnostic Displays (2500, 2520, 2550) . . . . . . . . . . . . . . . . . . . . . . 5.2
Diagnostic Displays (2510) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.3
General Fault Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.4
Serious Fault Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.6
General Spares List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1
Sample Cell Spares . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.2
Source Unit Spares . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3
Detector Spares . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3
Window Spares . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.4
Scrubber Spares . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.4
Recommended Spares . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.5
General Environmental Specifications . . . . . . . . . . . . . . . . . . . . . . . 7.1
Performance Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.2
Sampling Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.3
Sample and Ambient Influence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.3
Serial Output Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
Modbus RS-485 Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
02500 Safety Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.2
SECTION 1
1.1
INTRODUCTION
Introduction
This manual contains information regarding installation and hardware configuration of
the Servomex 2500 Series analysers (2500GEN/EU1/EU2/DU1/EX1/EX2/DX1,
2510GEN/EU1/EU2/DU1/EX1/EX2/DX1/HTV, 2520GEN/EU2/DU1/EX2/DX1,
2550GEN/EU1/EU2/DU1/EX1/EX2/DX1 variants).
GEN – General Purpose/Safe area use.
HTV – High Temperature Variant.
EU2/DU1 – Certified for European areas that require EPL Gc or Db.
EU1 – Certified for European areas that require EPL Gb.
EX2/DX1 – Certified for International areas that require EPL Gc or Db.
EX1 – Certified for International areas that require EPL Gb.
Note that the EU1, EU2, EX1, and EX2 versions can also be used where flammable
samples are to be monitored in otherwise safe areas.
Note: for more information regarding Equipment Protection Levels, see Appendix B.
The Servomex 2500 Series has been designed and manufactured using Servomex
quality systems, which have been certified to ISO9001.
A separate Quickstart manual is also supplied with the 2500, 2520 and 2550 analysers.
This details software configuration and operation of the analyser. An alternative
Quickstart manual is supplied with the 2510 analysers. Extra copies of either may be
ordered from Servomex.
Some variants are supplied with a Safety Certificate Manual that details the hazardous
area approvals and declarations applicable.
Details of the hardware and instructions for servicing, by qualified personnel only, are
presented in the 2500 Service Manual which may be ordered from Servomex.
Technical assistance and spare parts are available from Servomex outlets (or their
local agents) listed on the back cover.
Use this manual for:
Installation
To take commissioning to the point where the analyser is
powered and operational. The installer is advised to read
this manual completely before commencing installation.
Use the Quickstart manual for:
Configuration
How to set up the clock, passwords, alarm levels, analogue
outputs, relays and other parameters.
Calibration
How to use the manual and automatic calibration/checking
facilities.
Review
How to review configuration and set-up information.
1.1
1.2
General Description
The Servomex 2500 Series is a single beam, multi wavelength process analyser
suitable for monitoring up to three components in a gas sample stream (2550). It is
supplied configured to the customer's precise analytical requirements for a stated
analysis in a specific background stream.
The general layout of the 2500 Series is shown in Figure 1.1. It has been designed for
modern industrial environments with the emphasis on rugged construction, reliable
performance, simple operation and easy servicing. The analyser is controlled using an
on-board microprocessor, which gives flexibility to the user, enabling him/her to
configure the operation of the 2500 to suit the particular process environment. The
2500 is operated via a very simple control panel, which is mounted on the analyser
itself.
The gas or liquid sample to be analysed must be passed continuously through the
analyser's sample cell. The 2500 is designed for continuous 24 hour/day operation and
should not normally be switched off. Versions of the 2500 analyser are suitable for use
in safe and hazardous areas. They are also designed to I.P. 65. (without electrical cell
heater, I.P. 50 with cell heater).
Figure 1.1
2500 Overall View
Note that the EU1, EU2, EX1, and EX2 versions can also be used where flammable
samples are to be monitored in otherwise safe areas (see Safety Certificate Manual).
1.2
WARNING
The EU1 and EX1 versions are intended for use with a suitably certified Purge
Control System (PCS). Do not install the EU1 or EX1 version in a hazardous
area without a purge controller.
The analyser's software is menu driven and has been designed to be as intuitive as
possible, enabling the user to fully operate and control the 2500 with the minimum of
familiarisation.
NOTE
Full technical specifications for 2500 Series Analysers are presented in section 7 of this
manual.
1.3
Basic Construction
The basic construction of the 2500 is shown in Figure 1.2. It comprises 2 cast end
assemblies (with hinged opening/removable covers), connected by a rigid mounting
beam, or chassis. The sample cell is mounted between the 2 end assemblies and is
removable for cleaning. This configuration results in a single optical axis, i.e. straight
through beam design, with no alignment adjustments necessary. The source end (right
side) contains the soft U.V. or infrared source, which generates a broad beam of
energy across the appropriate spectrum. This is mounted on the chopper box
assembly, which contains the interference filters, mounted on a rotating chopper wheel.
These filters select the appropriate wavelengths for the analysis for which the 2500 is
configured. In most applications, there is one measurement wavelength and one
reference wavelength for each measured component.
The infrared or visible beam, alternating in wavelength, is then focused through the
sample cell, where energy at the measurement wavelength is selectively absorbed by
the components being analysed. The remaining energy falls on the detector in the
detector end.
The sample cell is a simple thick metal construction (316 stainless steel as standard)
fitted with optical windows (each 6 mm thick) to allow passage of the beam. It is
supported between the 2 ends of the 2500 by short cylindrical bosses, which may be
scrubbed or purged, depending on application. The whole cell assembly is easily
removed by loosening the two bosses and lifting it away, as described in Section 5.
1.3
Figure 1.2
2500 Construction
The detector end (left side) contains the detector assembly and also the main
electronic circuit boards. The display and control panel is also mounted in the hinged
lid of this end.
1.4
Options and Accessories
The 2500 supplied may have been fitted with customer specified options and/or
accessories. The full list of available items with brief descriptions is given here.
WARNING
The surface temperature of heated cells can reach or exceed 100°C
The following symbol is used on the heated cell:
CAUTION, HOT SURFACE
1.4.1
Heated Sample Cell – Electrical
Where fitted, this is powered and controlled by the 2500 itself and is suitable for gas
samples. It may be used in both general purpose and some hazardous areas. The
normal temperature setting for common gas applications is 60°C, although it may be
supplied set to any temperature from 60 – 180°C for HTV variants or from 60 – 130°C
for other versions.
1.4
1.4.2
Heated Sample Cell – Steam
Where fitted, this is a steam heating coil which must be supplied with low pressure
steam by the user and vented to atmosphere, to maintain a cell temperature of
approximately 100°C. This allows cell heating in hazardous areas up to Zone 1/Div. 1
classification, and is normally used with the EU1 and EX1 versions only.
1.4.3
Optional Output Cards
As standard, all 2500 Analysers have 2 analogue mA (current) outputs and 3 relay
contact pairs. The analogue outputs are user configured as described in the Quickstart
manual. The relay contacts are user-assigned to any combination of concentration
alarms, fault alarm, or calibration in progress, as similarly described in the Quickstart
manual. These standard outputs originate from the Sensor Interface Board (SIB PCB)
of the 2500.
An additional output PCB is available as option:
The 02000/916 PCB provides an extra 2 x (mA) plus 2 x (relay) outputs
This additional output PCB can be fitted into the normally vacant slot 2 in the 2500.
These extra outputs are configured along with the standard outputs in the usual way
(See Quickstart manual).
NOTE
Any or all of these analogue outputs may be configured as a voltage output by hardware
link selection, as an alternative to current output.
1.4.4
Compensation Options
The output of any infrared analyser is affected by changes in sample pressure and
sample temperature, unless compensation is applied. Alternatively, these parameters
can be kept constant by a suitable sample conditioning system. Ambient (barometric)
pressure will also affect sample pressure in the analyser if a gas sample is being simply
vented to atmosphere. The following compensation options may be fitted to a 2500.
•
Sample Pressure Compensation
A solid state pressure transducer is fitted and plumbed into one sample cell connection.
A continuous pressure measurement over the range of the standard transducer, 0 -345
kPa (0 – 50 psia), is made by the 2500. During calibration in the factory, an empirical
sample pressure/span concentration relationship is determined and stored in memory
to calibrate the compensation. Suitable for gas sample measurements only.
This option is detailed in Section 3.5.
1.5
WARNING
Standard transducers are not suitable for oxygen service.
•
Sample Temperature Compensation
A type 'K' thermocouple in stainless steel sheath is supplied, and a continuous
temperature measurement is made by the 2500 over the range -50 to 200°C (-58 to
392°F). During calibration in the factory, an empirical sample temperature/span
concentration relationship is determined and stored in memory to calibrate the
compensation. Suitable for gas or liquid samples, and for safe and hazardous area
installations. Not available with heated cell.
1.4.5
Digital Communications Options
As standard, all 2500 Analysers are equipped with an RS-232 port for the output of
configuration or measurement data to a printer, DCS or datalogger. As an alternative,
the analyser can be fitted with one of the following bi-directional digital
communications options:
•
Modbus ASCII or Modbus RTU over RS-485
•
Modbus TCP over Ethernet
When either of these options are fitted the standard RS-232 output is not available.
.
Appendix A describes how to access analyser data and control auto calibration using
the Modbus protocol.
1.4.6
Application Options
A range of standard and special application options is available with the 2500 and it will
have been supplied configured with one particular option to make the customerspecified measurements. Full details of the application configuration (i.e. optical
materials, operating wavelengths, cell path length, wetted materials etc.) will be found
on the Product Specification document in the front of this manual.
1.6
SECTION 2
2.1
INSTALLATION – GENERAL
Introduction
This section provides all the information required to install the 2500 analyser. The
installer is advised to read this section through completely before commencing the
installation.
2.2
Preparation
2.2.1
Tools
Installation of the 2500 requires the use of standard hand tools only, and an Allen key
provided with the instrument to open the end case covers. Preparation of suitable
mounting holes in support brackets or panels may require appropriate power tools.
2.2.2
Safety Precautions
The electrical power used in this equipment is at a voltage high enough to endanger
life therefore normal electrical safety precautions must be observed. Where necessary
a warning is placed prior to the procedure it refers to as shown in the following example:
WARNING
Always ensure that the electrical mains power supply is disconnected before
commencing work.
WARNING
Electrostatic hazard: Clean display and keypad using a clean cloth moistened
with water.
2.2.3
Unpacking and Inspection
WARNING
2500 analysers weigh in the range 25 kg (55 lbs) to 50 kg (110 lbs) and care
must be taken when handling. Use 2 people if necessary and only lift using
the yellow handles provided, not the sample cell.
Lift in the approved manner and remove the 2500 analyser from its packing and inspect
for any damage incurred during transit. If any damage has occurred inform Servomex
or their agents immediately. Retain packing, in case it is required to return the analyser
to the supplier.
2.1
After the initial visual inspection, carry out the following checks. Beware gas strut
action of the hinged covers.
•
Check that the serial number of the 2500 (located inside source end and also
marked on the external rating label) corresponds to the serial number given in
the Manufacturing Data Sheet found at the front of this manual.
•
Check that the specification sheet details are in order and agree with the
purchase requirements. Pay particular attention to the Manufacturing Data
Sheet supplied, and any Special Instrument Process Sheets attached.
•
Check that the accessories are present and undamaged. Standard 2500
accessories provided are: Fuse kit, Allen key, Plastic plugs (refer to section
2.4.2) and Ferrites (refer to section 3.2).
2.3
Electrical Connections
CAUTION
Specific voltage rating information is given on the label located on the outside of the
2500 Analyser and copied within the enclosure. Ensure that the analyser is correctly
configured for the supply voltage.
The 2500 Analyser has no integral switch for disconnecting the electrical supply.
The installer must include a means of isolating the electrical power by means of a
switch or circuit breaker located close to the 2500 Analyser. It must be marked as
the disconnecting device for the analyser and be easily accessible
The electrical supply to the 2500 Analyser MUST be protected by a suitable fuse or
over current protection device rated at no greater than 6 Amps.
Where the main electrical supply is not referenced to ground, an isolating transformer
must be fitted and its secondary winding suitably earthed. The following mains power
cable requirements must be met:
•
Three core cable with Line/Live (L), Neutral (N), Earth/Ground (E) conductors.
•
The cable voltage rating should conform to the supply voltage. The type of
cable used should conform to current local regulations, and be suitable for the
installation environment. For use in areas exposed to weather or mechanical
stress, armoured or sheathed cable is recommended.
•
All electrical cables enter the body of the 2500 Analyser via user-supplied cable
glands. These cable glands must be air tight and/or suitably approved if the
units are to be purged and/or mounted in hazardous areas.
•
The electrical supply connection terminals are suitable the following cable:
Flexible conductors
–
0.5 to 1.5mm2 (20 to 16 AWG)
Solid conductors
–
0.5 to 2.5mm2 (20 to 14 AWG)
Cables should be suitable for temperatures of at least 75°C.
2.2
2.4
Installation
WARNINGS
•
The 2500 Series analysers are not suitable for use in hazardous areas
unless the correct certification labels have been affixed by Servomex
and a suitable purge control system is installed (EU1 and EX1
versions).
•
The installer must be satisfied that the 2500 analyser installation
conforms to the relevant safety requirements, national electrical code
and any other local regulations, and that the installation is safe for any
extremes of conditions which may be experienced in the operating
environment of the analyser.
•
With the exception of relay contacts, Signal input and output terminals
of the analyser must only be connected to circuits that are separated
from mains voltage by at least reinforced insulation.
•
If the equipment is used in a manner not specified by the manufacturer
the protection provided by the equipment might be impaired.
•
Many process samples, gas streams and test samples are toxic,
asphyxiant, corrosive, flammable or a combination of some or all of
these. It is the responsibility of the installer to ensure that adequate
precautions are taken during installation of the 2500 analyser,
connection of samples, and during any other checks or tests. All
sample line connections must be checked for no leaks.
•
To avoid propagating brush discharges, the unit is not to be installed in
a high velocity dust-laden atmosphere.
•
The plastic cell cover (if fitted) may constitute and electrostatic
charging hazard. Use a damp (but not wet) cloth to wipe clean the outer
surface of the cover.
2.4.1
Positioning
Sensible positioning and an accurate and secure installation will minimise
maintenance, instrument breakdowns and will provide reliable service.
The location should be reasonably vibration free, subject to minimal fluctuations in
ambient temperature and provide reasonable access to the instrument and the
availability of any required services.
If the 2500 is to be mounted outside it must be suitably protected from all extremes of
climatic conditions by a suitable insulated enclosure, paying particular attention to the
ambient temperature range and the rate of temperature change specifications.
EU1 and EX1 versions must be used with a Purge Control System, which must meet
the necessary pressurisation requirements, and be supplied with the necessary
pressurisation and control components.
2.3
WARNING
EU1 and EX1 versions must be used with a Purge Control System, which
meets the necessary pressurisation and certification requirements as detailed
in the Safety Certificate Manual.
If a gas sample conditioning system is to be connected it must be designed to provide
a filtered, clean, non-condensing sample for the analyser with no risk of a possible
carry- over of condensate into the sample cell.
If a liquid sample conditioning system is to be connected it must be designed to provide
a filtered, single-phase sample for the analyser.
2.4.2
Mounting Details
The following paragraphs provide mounting information for the 2500 Analyser. Use the
yellow lifting handles to manoeuvre the 2500 into position, using two people where
necessary.
The analyser may be mounted vertically or horizontally, ideally at chest height for best
visibility and ease of access, but with some restrictions (refer to Fig 2.1).
CAUTION
•
For Liquids the analyser should be mounted horizontally with the cell
inlet port at the bottom.
•
For Gases the analyser should be mounted horizontally with the cell
inlet port on top, especially if there is risk of particulates or
condensates in the sample.
•
The 2500 may be mounted vertically for gas samples, provided the cell
outlet is at the bottom. This is only suitable if there is no risk of
particulates or condensates in the gas sample.
2.4
When mounting the analyser ensure that the panel or brackets employed are adequate
to take the weight and there is a minimum clearance above the casing of 500mm (20")
to provide space to open the end covers.
Fig 2.2 provides mounting details and a table of mounting hole spacings for the various
cell lengths. Be sure to identify the correct hole spacing.
When the 2500 is to be vertically mounted the display may be rotated 90° to bring it into
a legible position. This is done by removing the internal metal cover and ribbon cable
clamp, and removing the retaining bracket. The display unit can then be withdrawn a
little and rotated. Reassemble the disturbed components in reverse order, carefully
folding the ribbon cable to prevent any stress on it.
Figure 2.1
2500 Positioning Restrictions
2.5
Figure 2.2
2500 Mounting Details
Once mounted, the two yellow lifting handles and their spacers MUST be removed by
unscrewing the securing screws. Press the black pop-in plugs supplied into the vacant
screw holes and retain the handles for possible future use.
2.6
2.5
Instructions Specific to Hazardous Area Installations
(Reference European ATEX Directive 94/9/EC, Annex ll, 1.0.6.)
The following instructions apply to equipment covered by certificate numbers: Sira
09ATEX1341X, IECEx SIR09.0136X, Sira 10ATEX4015X, Sira 10ATEX 9016X and
IECEx SIR 10. 0007X
2.5.1
The equipment may be located where flammable gases and vapours of
groups llA, llB and llC may be present. The Ex n analyser may also be located
where combustible dusts of groups IIIA or IIIB may be present. The
equipment is only certified for use in ambient temperatures in the range -20°C
to +55°C for type 'p' approval and -10°C to +55 °C for Ex n approval, and
should not be used outside these ranges.
(Note performance specification is 0°C to +55°C)
2.5.2
All cable entries or closing devices for unused entries shall be suitably
certified to ensure that they comply with the general requirements of EN
60079-0:2009 and / or IEC 60079-0:2007.
2.5.3
For the IECEx certified analyser, all cable entries or closing devices must be
IECEx certified.
2.5.4
To prevent propagating brush discharges, the analyser must not be mounted
in a high velocity dust - laden atmosphere.
2.5.5
The equipment has not been assessed as a safety-related device (as
referred to by Directive 94/9/EC, Annex ll, clause 1.5).
2.5.6
Installation of this equipment shall be carried out by suitably trained
personnel in accordance with the applicable code of practice (EN 60079-14
within Europe).
2.5.7
Repair of this equipment shall be carried out by the manufacturer or in
accordance with the applicable code of practice (EN / IEC 60079-19).
2.5.8
If the equipment is likely to come into contact with aggressive substances,
then it is the responsibility of the user to take suitable precautions that prevent
it from being adversely affected, thus ensuring that the type of protection is
not compromised.
Aggressive substances
e.g. acid liquids or gases that may attack metals,
or solvents that may affect polymeric materials.
Suitable Precautions
e.g. regular checks as part of the routine
inspections or establishing from the material's
data sheet that it is resistant to specific
chemicals.
2.7
2.6
ATEX Label Information
2.6.1
EU1 T5 (Unheated Cell)
2.6.2
EU1 T4 (Steam Heated Cell)
2.8
2.6.3
EU2 T4 and DU1 T80°C (Unheated Cell)
2.6.4
EU2 T4 and DU1 T125°C (Heated Cell)
2.9
2.6.5
2.10
EU2 T3 and DU1 T175°C (Heated Cell)
SECTION 3
INSTALLATION – ELECTRICAL
WARNINGS
•
Lethal voltage: mains AC power supplies are potentially lethal. The
installer must ensure that the power supply has been isolated before
commencing installation.
•
The installer must be satisfied that the 2500 analyser installation
conforms to the relevant safety requirements, national electrical code
and any other local regulations, and that the installation is safe for any
extremes of conditions which may be experienced in the operating
environment of the analyser.
WARNING
The following symbols are used on the warning label fitted on the inside of the
Source End Enclosure Cover.
CAUTION, ELECTRIC SHOCK HAZARD
CAUTION, CONSULT MANUAL
3.1
Electrical Power Connections
For convenience the gas struts may be removed during installation by carefully levering
them off. Ensure that the cable used is suited to the installation environment, and is the
correct rating. Ensure that the fuse is suited to current rating of the analyser and that
all glands and cable connections are secured firmly.
The mains power connection to this instrument is made via an electrical filter unit
attached to the outside of the case. The power cable into this unit does not need to be
screened.
3.1
WARNING (EU1 AND EX1 VERSIONS)
AC Power for Zone 1 (EU1 and EX1) versions may only be connected via a
suitably certified Purge Control System. Refer to the PCS manual for details.
The mains power connection to the instrument is made via the purged
electrical filter unit attached to the outside of the case.
Refer to Figure 3.1 a) or b) for illustration.
To connect the 2500 analyser to the mains supply:
1.
Isolate the mains AC supply at source.
2.
Remove filter unit cover.
3.
Fit suitable cable gland to the mains filter unit.
4.
Insert cable through gland and secure.
5.
Wire to terminals.
6.
The mains AC supply is connected internally to TB30. The voltage selection is
made by wire link on TB31 as illustrated in Figure 3.1 a) or b).
7.
Connect the 2 external earthing studs to earth/ground. The RF earth provided
must be locally earthed using as suitable solid conductor connected to a local
physical earth to ensure optimum RFI protection. The external system earth must
be connected to the mounting panel / cubicle / frame / local Intrinsically Safe
earth, as appropriate.
DO NOT APPLY POWER YET.
NOTE
•
When installed in hazardous locations, appropriate wiring practice must be
used (e.g. within Europe installation must be in accordance with EN 6007914)
•
To ensure the correct function of the analyser the mains input shall be
protected from surges.
3.2
Figure 3.1a
2500 AC Power and Valve Connections (Purged Filter)
3.3
Figure 3.1b
3.4
2500 AC Power and Valve Connections (Non-Purged Filter)
Figure 3.2
2500 Series Terminal Locations
3.5
3.2
Signal Connections
NOTE
All signal cables must have a braided overall screen or armour. The screen must be
terminated at the point of entry to the case. This will be by using a gland, which
makes a connection between the cable screen and the case. Beware of ground/
earth loops if the screens are also connected at the user end.
Where two cables are passed through a single entry, a metal gland specifically
intended for two separate cables shall be used.
To minimize the effects of interference from RF fields, each mA output cable pair,
the RS485 cable (if fitted) and the Ethernet cable (if fitted) shall be looped once
through a Steward type 28B0562-200 or equivalent ferrite (Servomex Part No.
2824-0017). Refer to Figure 3.3.
Figure 3.3
3.6
Signal Cable Connections
3.2.1
Analogue Outputs
Figure 3.4
SIB and Optional Output PCB's
Each analogue output is supplied set for current output and may be configured by the
user to be 0-20mA or 4-20mA, and be assigned a particular range of operation, in
software. This is described in the Quickstart manual. In addition, a "Range 2" setting
for each analogue output can also be configured in software. Maximum impedance is
1k ohm for current output. If desired, each analogue output may be changed to voltage
output of either 0-10V or 2-10V, minimum impedance 1M ohms. This is done by
soldering links into position "LINK 10" and "LINK 11" on the relevant SIB (Sensor
Interface Board) PCB or Option PCB for the first and second analogue output
respectively. See Section 3.2.2. In every case, fit a suitable gland, insert, secure and
strip a suitable cable pair before connecting the cores to the appropriate terminals.
3.7
WARNING (EU1 and EX1 VERSIONS)
Where the 2500 is located in an area that requires EPL Gb, a suitable Relay Box
should be used to isolate the outputs. This Relay Box should be switched from the
Purge Control System.
Each analogue output (and each relay output) has an individual identity in software.
This identity is a number, which relates to the "slot" in which the PCB (SIB or Option
PCB) containing the actual output (or relay) physically resides.
For example, the first analogue output is from the standard SIB PCB which resides in
"slot 1" and therefore its software identity is "1.1" (i.e. first slot first output channel). The
second analogue output is "1.2", and the first relay output is "1.3", etc.
The following tables show the physical locations for connections to the analogue
outputs.
Table 3.1 – Analogue Output Connections
Output
Software ID
Analogue 1
1.1
TB 24-4 (+) and TB 24-3 (-)
Analogue 2
1.2
TB 24-2 (+) and TB 24-1 (-)
Analogue 3*
2.1
TB 22-1 (+) and TB 22-2 (-)
Analogue 4*
2.2
TB 22-3 (+) and TB 22-4 (-)
*option PCB
Terminal locations are as illustrated in Figure 3.2.
3.8
Terminal Connections
3.2.2
Analogue Output Link Selections on Sensor Interface Board (SIB) PCB
After connecting up the required analogue outputs, the user should make the following
hardware link selections on the SIB PCB and (where fitted) Option PCB(s). Refer to
Figure 3.3.4 for general schematic of both these types of PCB. Carefully remove the
PCB for link selection and be sure to replace it in the correct slot and in the correct
orientation.
•
Current/Voltage
Each output may be individually selected as current (0/4-20mA) or voltage
(0/2-10V). Factory setting is CURRENT.
Analogue o/p No.1
Analogue o/p No.2
•
Current:
Link 10 – OPEN
Voltage:
Link 10 – CLOSED (solder link)
Current:
Link 11 – OPEN
Voltage:
Link 11 – CLOSED (solder link)
Action on System Failure
To enhance fail-safe operation, each output may be selected to drive HIGH
(to 20.5mA/10.2V) or LOW (to 0 mA/0V.) if the 2500 detects a system failure
(Shutdown state, i.e. serious fault and measurement invalid). Factory setting is
HIGH.
Analogue o/p No.1
HIGH on Shutdown: SW1/3 OFF
LOW on Shutdown: SW1/3 ON
Analogue o/p No.2
HIGH on Shutdown: SW1/2 OFF
LOW on Shutdown: SW1/2 ON
•
Delay on Fail Safe Action
The detection of system failure (Shutdown, i.e. serious fault/measurement
invalid) is practically instantaneous and therefore a sufficiently large, rapid,
mains transient or brown-out may also cause the fail-safe action described
above to occur immediately. In installations where this would be undesirable,
the fail-safe action may be delayed for 5 seconds to overcome such possible
transient effects.
Factory setting is NO DELAY.
Both Analogue o/p
DELAY on fail-safe:
SW1/1 OFF
NO DELAY on fail-safe:
SW1/1 ON
3.9
3.2.3
Relay Outputs
NOTE
Each relay output is supplied set for NC (i.e. Normally Closed under "Safe"
condition) operation, and may be assigned to a particular alarm or signalling
function in software. This is described in the Quickstart manual.
User selection of NC or NO operation is described in Section 3.2.4. As with the
analogue outputs, each relay has its own software identity relating to the slot in which
the PCB, on which it is physically located, resides. The following tables show the
physical locations for connections to the relay outputs. All relay outputs are rated
240VAC/1.0A or 30VDC/1.0A. In each case, fit suitable glands and cable and connect
to the relevant terminals. Select the correct procedure according to the architecture.
Table 3.2 – Relay Output Connections
Output
Software ID
Terminal Connections
Relay 1
1.3
TB 23-1 and TB 23-2
Relay 2
1.4
TB 23-3 and TB 23-4
Relay 3
1.5
TB 23-5 and TB 23-6
Relay 4*
2.3
TB 22-5 and TB 22-6
Relay 5*
2.4
TB 22-7 and TB 23-7
*option PCB
Terminal locations are as illustrated in Figure 3.2.
3.2.4
Relay Output Link Selection
After connecting up the required relay output, the NC/NO selections may be altered on
the SIB PCB and (where fitted) output Option PCB(s). Refer to Figure 3.4 for general
schematic and carefully remove and replace the PCB as before, if a change is to be
made.
Normally Closed/Normally Open (NC/NO)
The factory setting of all relay output wire links is soldered in the NC position. This
means that the contacts will open to signal an alarm. The contacts will also be open if
there is no power applied to the 2500, i.e. they are fail-safe. If, however, it is required
to change to NO operation (i.e. close to signal an alarm) this can be done by cutting
and reordering the wire links as listed below.
3.10
WARNING
Setting the relays to Normally Open (N.O.) operation will mean that there is no
fail-safe action since any loss of cable continuity will prevent an alarm being
signalled.
NOTE
Regardless of N.O./ N.C. setting, the relays will always be in "alarm" state wherever
there is no power applied to the 2500, or when they are unassigned.
Relay o/p No. 1
N.C. Link 3 – A
N.O. Link 3 – B
Relay o/p No. 2
N.C. Link 4 – A
N.O. Link 4 – B
Relay o/p No. 3
N.C. Link 5 – A
N.O. Link 5 – B
3.3
Control Connections
NOTE
All control cables must have a braided overall screen or armour. The screen must be
terminated at the point of entry to the case. This will be by using a gland, which
makes a connection between the cable screen and the case. Beware of ground/earth
loops if the screens are also connected at the user end.
Where two cables are passed through a single entry, a metal gland specifically
intended for two separate cables shall be used.
As standard the following connections are provided on the 2500, and these are made
to the 2500 Analyser PCB's utilising a suitable gland in an adjacent hole. Refer to
Figure 3.2 and Table 3.3.
Table 3.3 – External Control Connections
PCB
Function
Connections
Transmitter PCB
02500904A
Range Change
TB26 – 1
TB25 – 3
Transmitter PCB
02500904A
Auto calibration Initialise
TB26 – 2
TB25 – 3
Transmitter PCB
02500904A
Password Keyswitch
TB26 – 3
TB25 – 3
Power Control PCB
02500911A
Flow Failure
TB36 – 1
TB36 – 2
3.11
3.3.1
Range Change Input
Every analogue output can be independently scaled in software to represent some or
all of the calibrated range of the instrument, as described in the Quickstart manual.
However, under some process conditions (e.g. plant start-up) it may be desirable to
have different output ranges on demand. When the Range Change Input is used all
analogue outputs will change to their user-designated "Range 2" settings. Note that
"Range 2" settings can be identical to "Range 1" settings if desired.
NOTE
"Range 2" is selected as long as contacts remain closed.
Range 2 settings selected by Contact closure between:
TB26 – 1 and TB25 – 3
3.3.2
Auto calibration Initiation Input
Auto calibration/Autocheck may be triggered by the user from the control panel or by
using the 2500's own real time clock system. Alternatively, it may be triggered at will
using a manual remote switch, by a host device closing a contact, or by means of a
Modbus command (where a digital communications option is fitted).
Auto calibration initialised by contact closure between:
TB26 – 2 and TB25 – 3
NOTE
•
Contacts must be closed for at least 2 seconds but no more than 59 seconds.
•
Remote initiation starts the "one cycle" routine using all the user-set
parameters. It will not repeat unless contact closure is made again.
•
User is advised to set the Auto calibration period in software to zero
(i.e. internal timer disabled) when using remote Auto calibration initiation,
to prevent duplication of Auto calibrations.
3.3.3
Password Keyswitch Input
As an alternative to the use of passwords via the control panel, a simple Keyswitch can
be connected which enables access to both Operator and Supervisor levels without
password demands. The user must first set SW1/8 to ON on the Microprocessor PCB,
to enable this feature. See the Quickstart Manual for more details.
Keyswitch setting selected by Contact closure between: TB26 – 3 and TB25 – 3
3.12
.
WARNING
If the Keyswitch is left "ON" (contacts closed), access is continuously
available to all normally protected areas of the software.
3.3.4
Sample Flow Sensor Input
If the sampling system supplying the process sample to the 2500 is fitted with a loss of
flow sensor, this can be connected to the 2500 to enable the "loss of sample flow"
diagnostic to operate, which in turn triggers the analyser's Fault alarm. See Figure 3.1.
2500 Power Control PCB 02501911A
Loss of flow diagnostic triggered by contact closure between: TB36 – 1 and TB36 – 2.
NOTE
In common with all fault diagnostics this alarm will only be cleared by restoration of
flow/opening of contacts.
WARNING
Where the 2500 analyser is in a hazardous area, suitable protection concepts
must be used to connect the loss of flow sensor to the 2500 analyser.
3.3.5
Solenoid Valve Drives
Auto calibration routines are provided as standard on the 2500, and if they are to be
used, suitable 3-way solenoid valves can be plumbed into the sampling system as
shown in Fig 3.5. Two or three valves may be used and all should be located
conveniently close to the 2500. Each valve is switched via the 2500 Power Control PCB
relays as illustrated. The valves are selected for operation by the 2500 as shown in
Table 3.4 and Table 3.5. Connect the valves to the appropriate terminals using suitable
screened cables and glands, and attach screens to the nearest earth stud with the
minimum length of screen. See Figure 3.1.
The Auto calibration relay contacts are rated at 240VAC/1.0A, 30VDC/1.0A. The
contacts are fitted with an R-C snubber (100R + 47nF) to provide transient protection
when used with AC solenoids. 220/240Vac solenoid valves with a power rating less
than about 6VA are not recommended because the 'leakage' current through the
snubber may prevent low power solenoids from de-energising correctly.
Solenoid valves may use an external power source, or the internal 24VDC/12VA power
source may be used. When using the internal power source, ensure that the 12VA
rating is not exceeded. DC solenoid valves must have built-in suppression diodes fitted
across the coil to protect the relay contacts.
3.13
WARNING
If installation is in a hazardous area, use suitable certified solenoid valves.
NOTE
Valve (SV3) reverts to OFF in Shutdown (Serious Fault) condition. The use of the
optional 3rd "Sample/Inert valve will prevent sample entering the 2500 sample cell
while the cell or chopper box under temperature diagnostic is activated, i.e. heated
cell is not fully warmed up. This can be used whether or not Auto calibration is used
and will be appropriate in some applications: e.g.
•
if there is a possibility of condensation forming in the sample cell during warm
up, then nitrogen or another purge gas must be used to flush the cell during
it's warming up period.
•
if a hazardous sample is being monitored, the cell should be purged in the
event of a serious fault.
3.3.6
Externally Powered Solenoid Valve Connections
Connect via solenoid valve relays as detailed. Ensure that the relay rating is not
exceeded.
Table 3.4 – Externally Powered Solenoid Valve Connections
Valve State (ON=Current Flow Through
Solenoid)
3.14
Span/Zero
Valve SV1
TB34 – 1
TB34 – 2
Cal/Sample
Valve SV2
TB34 – 3
TB34 – 4
Sample/Inert
Valve SV3
TB34 – 5
TB34 – 6
Cell or Chopper Box
Under Temperature
either
either
OFF(Inert)
NORMAL STATE OFF
OFF (Zero)
ON(Sample)
ON(Sample)
Zero Sample Required
OFF (Zero)
ON(Cal.)
ON(Sample)
Span Sample Required
ON(Span)
ON(Cal.)
ON(Sample)
3.3.7
Internally Powered Solenoid Valve Connections
Connect to solenoid valve relays SV1, SV2 and SV3 as detailed below. Take care to
ensure that internal 12VA (total) rating of the 24VDC supply is not exceeded.
Table 3.5 – Internally Powered Solenoid Valve Connections
Valve State (ON=Current Flow Through
Solenoid)
Span/Zero
Valve SV1
TB34 – 1 (-)
TB36 – 6 (+)
Cal/Sample
Valve SV2
TB34 – 3 (-)
TB36 – 5 (+)
Sample/Inert
Valve SV3
TB34 – 5 (-)
TB36 – 5(+)
Cell or Chopper Box
Under Temperature
either
either
OFF(Inert)
NORMAL STATE OFF
OFF (Zero)
ON(Sample)
ON(Sample)
Zero Sample Required
OFF (Zero)
ON(Cal.)
ON(Sample)
Span Sample Required
ON(Span)
ON(Cal.)
ON(Sample)
NOTE
To ensure internally powered solenoid valve function, fit links between:
TB34 – 2 and TB36 – 4
TB34 – 4 and TB36 – 3
TB34 – 6 and TB36 – 3
CAUTION
TB36 terminals 5 and 6 (+24V) are always live and should not be grounded.
3.15
Figure 3.5
3.4
Auto calibration Valve Configuration
Digital Connections
NOTE
All digital cables must have a braided overall screen or armour. The screen must be
terminated at the point of entry to the case. This will be by using a gland, which
makes a connection between the cable screen and the case. Beware of ground/earth
loops if the screens are also connected at the user end.
Where two cables are passed through a single entry, a metal gland specifically
intended for two separate cables shall be used.
Digital connections in RS/EIA-232 format are provided for attaching a local portable
printer to receive set-up information directly down loaded from the 2500, or to report
measurement and analyser status information to a DCS or data logger.
3.4.1
RS-232 Connection
See the Quickstart manual for configuration of the RS-232 port and operation of this
function. See Figure 3.2 and Table 3.9 for connections. The format of the serial data
stream is as follows:
3.16
Table 3.6 – Serial Data Format
Item
Size
Description
1 byte
Carriage return character (ASCII code 13).
date;
8 bytes
dd/mm/yy or mm/dd/yy depending on user
settings.
time;
8 bytes
hh:mm:ss
number of
components
1 byte
1 byte Number of components fitted
(range 1-3)
2 bytes
Carriage return and Line Feed characters
(ASCII codes 13 and 10 respectively).
component 1 formula;
Max 6 bytes
Chemical formula defined for component 1.
component 1
concentration;
Max 5 bytes
Concentration for component 1 as defined
for measure display.
component 1 units;
Max 3 bytes
Units defined for component 1.
component 1 alarm
status;
4 bytes
One byte for each alarm. Set to alarm
number (1,2,3 or 4) when alarm is raised,
when alarm not raised
component 1 Auto
calibration status;
1 byte or
15 bytes
Indicates Auto calibration phase.
0 = Not in Auto calibration
1 = In span preflush
2 = In zero cal
3 = zero corrected. This is then followed by
two additional values, which are separated
by commas, each 6 bytes that represent the
zero before calibration and the zero after
calibration respectively.
4 = In span cal
5 = span corrected. This is then followed by
two additional values, which are separated
by commas, each 6 bytes that represent the
span before calibration and the span after
calibration respectively.
6 = In post flush
2 bytes
Carriage return and Line Feed characters
(ASCII codes 13 and 10 respectively).
Note: The data for components 2 and 3 will only be present on a 2550
instrument.
component 2 formula;
Max 6 bytes
Chemical formula defined for component 2.
(if fitted)
3.17
Table 3.6 – Serial Data Format
component 2
concentration;
Max5 bytes
Concentration for component 2 as defined
for measure display. (if fitted)
component 2 units;
Max 3 bytes
Units defined for component 2. (if fitted)
component 2 alarm
status;
4 bytes
One byte for each alarm. Set to alarm
number (1,2,3 or 4) when alarm is raised,
when alarm not raised. (if fitted)
component 2 Auto
calibration status;
1 byte or
15 bytes
Indicates Auto calibration phase.
0 = Not in Auto calibration l
1 = In span preflush
2 = In zero cal
3 = zero corrected. This is then followed by
two additional values, which are separated
by commas, each 6 bytes that represent the
zero before calibration and the zero after
calibration respectively.
4 = In span cal
5 = span corrected. This is then followed by
two additional values, which are separated
by commas, each 6 bytes that represent the
span before calibration and the span after
calibration respectively.
6 = In post flush
2 bytes
Carriage return and Line Feed characters
(ASCII codes 13 and 10 respectively).
component 3 formula;
Max 6 bytes
Chemical formula defined for component 3.
(if fitted)
component 3
concentration;
Max 5 bytes
Concentration for component 3 as defined
for measure display. (if fitted)
component 3 units;
Max 3 bytes
Units defined for component 3. (if fitted)
component 3 alarm
status;
4 bytes
One byte for each alarm. Set to alarm
number (1,2,3 or 4) when alarm is raised,
when alarm not raised. (if fitted)
3.18
Table 3.6 – Serial Data Format
component 3 Auto
calibration status;
1 byte or
15 bytes
Indicates Auto calibration phase.
0 = Not in Auto calibration
1 = In span preflush
2 = In zero cal
3 = zero corrected. This is then followed by
two additional values, which are separated
by commas, each 6 bytes that represent the
zero before calibration and the zero after
calibration respectively.
4 = In span cal
5 = span corrected. This is then followed by
two additional values, which are separated
by commas, each 6 bytes that represent the
span before calibration and the span after
calibration respectively.
6 = In post flush
2 bytes
Carriage return and Line Feed characters
(ASCII codes 13 and 10 respectively).
Note: End of 2550 specific data
fault status;
1 byte
fault numbers;
Set to 'F' when fault is active, when
no fault active
List of fault numbers of active faults,
separated by ','. Fault numbers are defined
in table 3.7.
2 bytes
Carriage return and Line Feed characters
(ASCII codes 13 and 10 respectively).
Chopper
temperature; ºC;
Max 4 bytes
Chopper temperature as displayed in
diagnostic display.
Compensation
temperature; ºC;
Max 4 bytes
Compensation (detector) temperature as
displayed in diagnostic display.
cell/sample
temperature; ºC;
Max 5 bytes
Cell/sample temperature as displayed in
diagnostic display.
source voltage; V;
Max 5 bytes
Source voltage as displayed in diagnostic
display.
Intensity 1;Intensity 2;
Intensity 3;Intensity 4;
Intensity 5;Intensity 6;
Max 6*6
bytes
Intensity channel values – diagnostic values
representing up to six input positions used
for measure and reference signals.
2 bytes
Carriage return and Line Feed characters
(ASCII codes 13 and 10 respectively).
3.19
Table 3.6 – Serial Data Format
checksum;
4 bytes
ASCII representation of 16 bit checksum
(modulo 65536), calculated by adding all
data preceding the checksum together.
2 bytes
Carriage return and Line Feed characters
(ASCII codes 13 and 10 respectively).
2 bytes
Carriage return and Line Feed characters
(ASCII codes 13 and 10 respectively).
Serial port parameters are as follows:
Baud rate: 2400
Parity: Even
Stop Bits: 1
Data Bits: 8
Handshake: hardware
Example Data:
Note: Checksums have not been calculated
Normal Operation Example Output
09/01/99;13:42:30;2;CO;0.5;%; 2 4;0;CO2;3.8;%; ;0; ;;50.0;51.5;183.2;5.0;
123456;521452;319322;4235421;478525;368965;????;
3.20
Table 3.7 –
Fault Number
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
Fault Number Reference
Display Message
OPTICAL BENCH POWER FAILURE
SOURCE VOLTAGE HIGH
SOURCE VOLTAGE LOW
INFRARED SOURCE FAILURE
CHOPPER MOTOR OUT OF LOCK
SAMPLE FLOW FAILURE
CHOPPER TEMP HIGH
CHOPPER TEMP LOW
CHOPPER TEMP SENSOR FAILURE
CELL TEMP HIGH
CELL TEMP LOW
CELL TEMP SENSOR FAILURE
SAMPLE TEMP SENSOR FAILURE
PRESSURE SENSOR FAILURE
BAD REFERENCE VOLTAGE
COMPENSATION T SENSOR FAILURE
DETECTOR SIGNAL HIGH
DETECTOR SIGNAL LOW
Reserved
Reserved
Reserved
AUTO CAL BAD PREFLUSH
ZERO OUT OF TOLERANCE
SPAN OUT OF TOLERANCE
Reserved
Reserved
Reserved
Reserved
Reserved
PASSWORD VIOLATION
SYSTEM CLOCK INOPERATIVE
SYSTEM ADC INOPERATIVE
DEFAULT SYSTEM DATA CORRUPTION
CALIBRATION DATA CORRUPTION
UTILITY DATA CORRUPTION
ALARM/RELAY DATA CORRUPTION
ANALOGUE ASSIGN DATA CORRUPTION
BAD SPAN IGNORED
3.21
Some characters defined by the 2500 are non-standard characters, i.e. not part of the
ASCII character set. These either occupy the ASCII codes above 127 or redefine an
ASCII code as another character. When these characters are used, for example in the
component formula, they will be substituted with a valid ASCII character prior to output
on the RS232. The non-standard characters and their substitute values are as follows:
Table 3.8 – ASCII Characters
ASCII Code
Character
Substitute Character
126
?
127
?
128
0
0
129
1
1
130
2
2
131
3
3
132
4
4
133
5
5
134
6
6
135
7
7
136
8
8
137
9
9
138
?
139
?
140
141
??
142
°
143
£
Table 3.9 – RS232 Connections
2500 Terminals
Connector
External RS232 Terminals
0V (Ref.)(0VR)
TB20 – 1
Signal GND/0V
Clear to Send (CTS)
TB20 – 3
Data Terminal Ready (DTR)
Transmit (Tx)
TB20 – 4
Receive (Rx)
3.22
WARNING
Do not use an uncertified dcs, datalogger or printer in a hazardous area.
3.4.2
Modbus RS-485 Connection
Modbus communications over an RS-485 interface requires the Communications
Board RS-485 option PCB 02500912. This is fitted vertically above the Power Control
PCB in the source end assembly.
When connected to a Modbus network the analyser may be an end-point node on a
single cable (Cable 1), or it may be a mid-point node with two cables (Cable 1 and
Cable 2). These connections are made as shown in Table 3.10.
Table 3.10 – Modbus RS-485 Connections
Field connections to PCB 02500912
Connecting Cable
TB3 Pin 1
“A” from Cable 1
TB3 Pin 2
“B” from Cable 1
TB3 Pin 3
“A” from Cable 2 (if fitted)
TB3 Pin 4
“B” from Cable 2 (if fitted)
Note that if Cable 2 is not fitted, a link must be made betweenTB3 Pin 4 and TB3 Pin
5 to provide the correct cable termination. Cables must be looped through a ferrite (see
section 3.2)
See the Quickstart manual for the selection and configuration of Modbus
communications.
3.4.3
Modbus TCP Ethernet Connection
Modbus TCP communications over Ethernet requires the Communications Board
Ethernet option PCB 02500913. This is fitted vertically above the Power Control PCB
in the source end assembly.
Ethernet cabling should use shielded (STP) CAT 5 cable or better.
The cable should first be passed through the appropriate gland and looped through a
ferrite (see section 3.2) and then assembled onto a standard RJ45 connector. The
cable screen/shield must be terminated as instructed. The table below list the functions
of each connector pin.
Pin 1
Transmit +
Pin 2
Transmit -
Pin 3
Receive +
Pin 6
Receive -
Pin 4, 5, 7, 8
Not used
3.23
For guidance, the tables below show the corresponding wire colours specified in TIA/
EIA-568A and TIA/IEA-568B for patch leads as viewed from the back of the plug with
the tab facing down. It is the responsibility of the installer to ensure the correct
connections are made.
Table 3.11 – TIA/EIA-568A Wiring
1
2
3
4
White/Green
Green
White/Orange
5
6
7
8
7
8
Orange
Table 3.12 – TIA/EIA-568B Wiring
1
2
3
White/Orange
Orange
White/Green
4
5
6
Green
See the Quickstart manual for the selection and configuration of Ethernet
Communication.
3.5
Pressure Transducer Connections – (If Supplied)
The pressure transducer cabling is terminated to the source end right hand side (RHS)
of the analyser.
The pressure transducer is supplied pre-wired to the analyser. The cabling is
connected to PCB 02500911A TB35 as shown in Table 3.13
.
Table 3.13 –
PCB 02500911A
Pressure Transducer Connections
Interconnecting Cable
Pressure Transducer
TB35 – 5 (signal +)
Identified '1'
Terminal 1 (+)
TB35 – 1 (signal -)
Identified '2'
Terminal 2 (-)
The general layout of the pressure transducer option is illustrated in Figure 3.6.
3.24
Figure 3.6
Typical Pressure Transducer Assembly
3.25
3.26
SECTION 4
4.1
INSTALLATION GAS CONNECTIONS
Purge Connections (EU1/EX1)
The next phase of the installation is to connect up any safety purge system required for
safe operation in a hazardous area.
4.1.1
General Purge Connections
WARNING (EU1 AND EX1 VERSIONS)
Where appropriate, the user must refer to the detailed instructions provided
in the manual supplied with the Purge Control System, and use them in
conjunction with the following instructions.
OPTICAL BENCH
60VA
60VA
Sample
Cell
Incandescent Particles
not Normally Produced
2.825 Litres
Maximum Volume
11.30 Litres
No Internal Source
of Release
Protective Device Showing
Automatic Disconnection of
Both Poles of Signal & Alarm
Circuits, 265V max
Maximum Volume
11.30 Litres
See Note C) & Note 1)
Inert Gas/Air Supply
Electrical Mains
Filter
1.10 Litres
Power 265V max
See Note C) & Note 1)
Purge Control System
ATEX Certified to EN60079-2:2007
or
IECEx Certified to IEC 60079-2:2007
Purge Exit
HAZARDOUS AREA
Figure 4.1
SAFE AREA
Purge Connection Schematic
4.1
The general schematic for a hazardous area purge system is shown in Figure 4.1.
WARNINGS (FIGURE 4.1)
1.
If this component is mounted in a hazardous area it should be protected
by one of the types of protection listed in EN/IEC 60079-0.
2.
All cable glands for use in electrical cable entries shall be of a type which
seals and clamps the cable.
NOTES (FIGURE 4.1)
a
The purge control system (PCS) shall be suitably certified to EN/IEC 600792:2007 and shall have the following characteristics:
(Note: if the 2500 series equipment is affixed with the IECEx label, then the
PCS unit must be IECEx certified):
Minimum permitted purge rate 10 L/min
Minimum permitted purge duration 11 minutes
Minimum overpressure 0.5 mbar (50 Pa)
Maximium overpressure 80 mbar (8000 Pa)
b
All monitoring shall be carried out in the return leg.
c
The PCS shall include an automatic device that operates when overpressure
falls below the minimum prescribed value. The device shall operate a pressure
switch that can be used for an alarm and/or isolation purposes.
d
When the PCS is used to protect the series combination of enclosures, the
protective gas supply shall be connected from the PCS to the first enclosure
and serially connected to the remaining enclosures.
e
Overpressure shall be monitored by the PCS unit to ensure that the minimum
overpressure is maintained and the maximum overpressure specified (80
mbar) is not exceeded.
f
When the certified PCS unit is located in an area that requires EPL Gb and if
the certified PCS unit is not fitted with a normally closed relief vent, a protective
device shall be fitted to prevent sparks or incandescent particles from being
ejected.
g
The PCS is required to carry out a purge cycle after a pressure loss before
power can be restored to the equipment.
h
The PCS shall contain a pressure sensing element which is factory set to a
level not less than the specified minimum (50 Pa, 0.5 mbar) and shall measure
pressure at the point in the system where the lowest pressure occurs.
i
The PCS shall supply a flow rate of protective gas not less than 10 standard
litres per minute and shall incorporate an automatic device to monitor flow
during the purge time specified below.
j
The PCS shall maintain the minimum specified overpressure during the purge
phase and afterwards.
4.2
NOTES (FIGURE 4.1)
k
The PCS shall incorporate a purge timer that shall be set to not less than the
minimum specified (11 minutes). The purge timer shall be re-set to zero if either
the minimum overpressure or minimum purge flow is not maintained.
l
The PCS shall incorporate a device to ensure that power is not connected to
the apparatus being protected until the specified purge time has elapsed.
4.2
End Boss Purge Connection
Depending on application, fittings for 1/8" o.d. tube may be provided to purge the end
bosses on either side of the sample cell and these should now be connected. Refer to
the following application notes and Figure 4.2 where necessary.
4.2.1
Introduction
The 2500 measuring cell is mounted on two short tubes or "end bosses" which
preserve the integrity of the optical system, while providing thermal isolation. The
measuring cell and the analyser enclosure are both independently sealed using
windows or lenses and elastomeric seals. The end bosses form the interface between
the two, and the implications of a cell window leak must be considered when
flammable, corrosive or toxic gases are measured.
Under ideal circumstances, the gas in the end bosses should be totally transparent to
the wavelengths of infrared energy used for the measurement. However, simply filling
the bosses with nitrogen and sealing them is not always satisfactory, because the
elastomeric seals are permeable and atmospheric gases such as carbon dioxide or
water vapour can leak in. Specifically, a problem will arise when the gas being
measured is also carbon dioxide or water vapour, or when the measurement is cross
sensitive to either of those gases: the effect of a gradual ingress of the atmospheric
gases will be seen as drift in the measurement. A further problem arises with low-level
measurements of carbon monoxide. In this case it is out-gassing of carbon monoxide
from the seals themselves which can cause drift.
4.2.2
End Boss Scrubbers
Problems, relating to cross interference within the end bosses, can usually be resolved
by fitting chemical "scrubbers" to the end bosses. These are small metal housings
packed with a chemically active material, which absorbs or reacts with the problem gas.
A number of materials are available which absorb water vapour (molecular sieve),
carbon dioxide (soda lime) and carbon monoxide (Hopcalite).
When a 2500 is specified, a suitable scrubber material is stipulated for the end bosses,
where necessary, and the analyser will be delivered with these scrubbers fitted.
NOTE
If the analyser is dismantled for any reason, the scrubbers must not be left exposed to
normal atmosphere for more than a few minutes or they will be exhausted. Similarly, after
long-term use they may need to be replaced, and this must always be with the correct type.
4.3
4.2.3
End Boss Vents
When a flammable, severely corrosive or toxic gas or liquid is present in the
sample cell, either as the sample or a background gas, additional issues arise.
The construction of the cell window seal is such that any leak of the seal or failure of
the window will result in the sample gas entering the end boss, but not being released
to the surrounding atmosphere or directly into the analyser enclosure. This is a
deliberate safety feature of the design. However it is necessary that steps are taken to
deal with the leak so that no permanent hazard arises, and permanent sealing of the
end bosses and the use of scrubbers is not generally acceptable. Instead, the end
bosses will be supplied with adaptors for connecting vent or purge lines.
If the measurement is not affected by atmospheric gases, then a single vent connection
of at least 4mm I.D. should be made to each end boss, which will vent to atmosphere
in a safe location where release of the sample will not cause a hazard. If the sample
pressure exceeds 50kPa (7 psig), the vent should be 8mm I.D., and the sample cell
outlet connection should be fitted with a non return valve, and the sample cell inlet
connection should be fitted with a restrictor which will limit the total flow to less than
10 litres/min in the event of total window failure.
4.2.4
End Boss Purge
If the measurement is affected by atmospheric gases, then in addition to the vent
described above, the second connection to each end boss should be connected to a
supply of dry (1) nitrogen (2) regulated to a flow of 100ml/min using a suitable flow
meter. Ensure that the end boss purge is always vented to atmospheric pressure and
the pressure within the end boss dead volume space does not exceed 2.5psig.
Suitable precautions should be made to ensure that, in the unlikely event of a
catastrophic failure of the cell window /sealing, sample gas is prevented from
contaminating the purge gas supply.
(1)
Dry means a dew point of below -30 deg C.
(2)
Dry CO2-free air may also be used in most applications except low-level CO2.
Hazardous Area Certification – flammable samples, additional note
The analysers are certified for use in either Zone 2 or Zone 1. In all cases this
certification does not permit the analyser (or any part of its enclosure) to be exposed to
a permanently flammable atmosphere. Sealing the end-bosses with scrubbers when
measuring a flammable sample would mean that should there be a cell window leak,
the part of the enclosure that the end boss seals against would become exposed to a
permanently flammable atmosphere. Venting or purging the end bosses resolves this
issue.
4.4
NOTE
Vent lines, if required, should be connected to end boss bottom tappings. Purge gas
should be connected to top tappings when required. When not required, top tappings
should be plugged.
IF FITTED, COMMENCE END BOSS PURGE FROM THIS POINT ONWARDS.
WARNING
Where the sample is corrosive or toxic, the vent lines from the end bosses
must take away any potential leakage to a safe disposal point.
Figure 4.2
Cell End Boss Connection
4.5
4.3
Steam Heated Cell
Some 2500's are fitted with steam heated cells, and at this stage suitable low pressure
steam, typically at 14-21kPa, 2-3 psig venting to atmosphere after the cell, should be
connected ready for use, but do not pass steam into the heating coil yet. Fittings for
1/8" o.d. tube are normally provided.
4.4
Process Stream Connection
4.4.1
Basic Operation
The process stream connections to and from the 2500 sample cell inlet and outlet
should now be made. The inlet and outlet are 1/4" o.d. stubs. Connections will normally
be from a local sample conditioning system. Ensure that any pressure compensation
transducer is correctly fitted. Ensure that all connections are made securely with the
correct compression fittings, of suitable materials. Gas samples should enter the cell
from above. (Refer to Figure 2.1).
Do NOT pass process samples into the 2500 yet.
4.4.2
High Integrity Operation
This is the recommended mode of operation and where appropriate, the process
sample can be routed via the "Sample/Inert" solenoid valve, controlled by the 2500,
which will only permit sample to enter the sample cell when it is fully warmed up and
no serious faults are present. See Section 3.3.5.
Do NOT pass process samples into the 2500 yet.
WARNING
In either case for toxic/flammable samples ensure that the sample inlet line
has a suitable restrictor inserted to limit sample flow in case of catastrophic
failure of sample tubing/cell connections.
4.5
Power-Up
1.
Confirm that all electrical and plumbing connections have been correctly made as
described in the preceding sections.
2.
Ensure purge gas is applied to the Purge Controller (Where Fitted).
3.
Apply power to the 2500 (via the Purge Control System for EU1 & EX1 Versions).
4.
Apply steam to any steam-heated cells fitted.
5.
The 2500 will show the start-up and Identity display as follows:
– Self-checks (Fault and Alarm LEDs tested)
– System O.K.
– Analyser identity/Software revision number.
– Serial number and order number.
– Calibration (measurand, range, units).
– Commence measurement.
4.6
6.
The normal measurement display will then appear and the 2500 will be
operational.
NOTES
•
The measurement will not be valid at this point since process sample is not
yet flowing.
•
If a heated cell is fitted, it will not yet be up to temperature and the intermittent
"warming up" message will be displayed. Depending on set point and/or
environment this may take several hours to reach temperature and the
message to cease.
•
The 2500 chopper box will also take up to 2 hours to reach specified
temperature and the "warming up" message will also be displayed until this is
accomplished.
WARNING
Check for no leaks of sample at full operating pressure and temperature
before proceeding. This is especially important for toxic/flammable samples.
While the analyser is warming up, the software configuration to suit the user's particular
requirements may now be accomplished. Proceed to the Quickstart Manual for details
of configuration. Note that during the warming up period, all outputs and alarms are live
and fully operational. All diagnostics are also operational, EXCEPT cell temperature
and chopper box temperature. These last two diagnostics only become operational
once the temperature set points have been reached or after 2 hours (for chopper box)
or 10 hours (for cell) have elapsed from power-up.
At this point process sample may be turned on and allowed to flow through the
2500 sampling system.
NOTE
Where the 3rd (sample/inert) solenoid valve is in use (see Section 3.3.5) the inert
medium will automatically be supplied to the analyser until the cell reaches correct
temperature.
4.7
4.8
SECTION 5
5.1
FAULT DIAGNOSIS AND CELL MAINTENANCE
Introduction
A fault will be signalled wherever any parameter measured by the 2500's diagnostics
becomes out of tolerance. Any Fault condition will be indicated by the Fault LED on the
control panel being illuminated.
The Fault signal can be assigned to a Relay, as described in the Quickstart Manual.
Faults are categorised as either:
1.
General Faults – parameter out of tolerance, intervention is required as
soon as possible. The analyser is still operational and measuring.
2.
Serious Faults (Shutdown) – performance seriously degraded,
measurement withdrawn and analogue outputs driven HIGH (or low, as
configured in Section 3.2.2).
NOTE
For detailed instructions on all Service and Maintenance operations for suitably
skilled personnel, please refer to the 2500 Series Service Manual.
5.2
Diagnostics
The diagnostics listed in Table 5.1 (2500, 2520 & 2550) and in Table 5.2 (2510) can be
interrogated via the DIAGNOSTICS function in DISPLAY menu. See Quickstart
Manual.
5.1
Table 5.1 – Diagnostic Displays (2500, 2520, 2550)
Display
Description
CHOPPER TEMP °C
Chopper box temperature
COMPENSATION °C
Detector temperature
CELL TEMP °C
Cell temperature (if enabled)
SAMPLE TEMP °C
Sample Temperature *
[Measurement 1] ABS au
Absolute absorbance units (M1)
[Measurement 2] ABS au
Absolute absorbance units (M2) *
[Measurement 3] ABS au
Absolute absorbance units (M3) *
SAMPLE PRESSURE (units)
Sample pressure *
SOURCE VOLTS V
Source voltage
MOTOR DRIVE %
SET POINT %
Motor drive power and set point
DETECTOR SIG M1
Detector signal level (M1) measurement
DETECTOR SIG M2
Detector signal level (M2) measurement *
DETECTOR SIG M3
Detector signal level (M3) measurement *
DETECTOR SIG R1
Detector signal (R1) reference
DETECTOR SIG R2
Detector signal (R2) reference *
DETECTOR SIG R3
Detector signal (R3) reference *
TEMP CORR M1
Temperature Correction (M1)
TEMP CORR M2
Temperature Correction (M2) *
TEMP CORR M3
Temperature Correction (M3) *
TEMP CORR R1
Temperature Correction (R1)
TEMP CORR R2
Temperature Correction (R2) *
TEMP CORR R3
Temperature Correction (R3) *
* (IF FITTED).
5.2
Table 5.2 – Diagnostic Displays (2510)
Display
Description
CHOPPER TEMP °C
Chopper box temperature
COMPENSATION TEMP °C
Detector temperature
CELL TEMP °C
Cell temperature (if enabled)
SAMPLE TEMP °C
Sample temperature (if enabled)
SAMPLE PRESSURE (units)
Sample pressure (if fitted)
SOURCE VOLTS
Source voltage
ABSORBANCE au
Absolute absorbance units
MOTOR DRIVE %
SET POINT %
Motor drive power and set point
DIFFERENCE SIG
Difference signal
DETECTOR SIG N
Detector signal (nitrogen)
DETECTOR SIG G
Detector signal (gas)
TEMP CORR D
Temperature correction (difference signal)
TEMP CORR N
Temperature correction (nitrogen)
TEMP CORR G
Temperature correction (gas)
5.3
5.3
General Fault Conditions
When the Fault LED is illuminated, the DISPLAY FAULTS function should be used to
determine the nature of the fault. The diagnostics (Section 5.2) may assist in defining
the cause, together with the comments given in Table 5.3. Action to cure the fault
should be taken as soon as possible.
Table 5.3 – General Fault Messages
Display Message
History Log
Entry
SOURCE VOLTAGE
HIGH
SOURCE HIGH
Possible Power Control PCB fault
SOURCE VOLTAGE
LOW
SOURCE LOW
Possible Power Control PCB fault
SAMPLE FLOW
FAILURE
FLOW FAIL
User Sample Flow Alarm active
CHOPPER TEMP HIGH
CH TEMP HIGH
Ambient temp. too high/control
problem
CHOPPER TEMP LOW
CH TEMP LOW
Ambient temp. too low/control
problem
CHOPPER TEMP
SENSOR FAILURE
CH TEMP FAIL
Chopper sensor failed/
disconnected/outside range.
CELL TEMP HIGH
CELL T HIGH
Possible Power Control PCB fault
CELL TEMP LOW
CELL T LOW
Possible Power Control PCB fault
CELL TEMP SENSOR
FAILURE
CELL T FAIL
Cell sensor failed/disconnected/
outside range
SAMPLE TEMP
SENSOR FAILURE
SAMPLE T FAIL
Sample sensor failed/disconnected/
outside range
PRESSURE SENSOR
FAILURE
NO PRESSURE
Sample sensor failed/disconnected/
outside range
BAD REFERENCE
VOLTAGE
REF OLT FAIL
Possible Transmitter PCB fault
DETECTOR SIGNAL
LOW
DET SIG LO
Probable cell contamination – clean
cell
COMPENSATION T
SENSOR FAILURE
COMP T FAIL
Detector sensor failed/disconnected
ZERO OUT OF
TOLERANCE
ZERO OUT TOL
Failed Autocalibration – bad zero
sample – use One Cycle or Manual
Zero
5.4
Description/comments
Table 5.3 – General Fault Messages
SPAN OUT OF
TOLERANCE
SPAN OUT TOL
Failed Autocalibration – bad span
sample – use One Cycle or Manual
Span
AUTOCAL BAD
PREFLUSH
BAD PREFLUSH
Failed Autocalibration – bad
preflush – use One Cycle
PASSWORD
VIOLATION
BAD PASSWORD
3 consecutive Password failures
SYSTEM CLOCK
INOPERATIVE
BAD CLOCK
Reset time or replace Clock chip
DEFAULT SYSTEM
DATA CORRUPTION
BAD SYS DATA
Factory calibration corrupt – use
Associate
SYSTEM ADC
INOPERATIVE
BAD ADC
Possible SIB PCB fault
CALIBRATION DATA
CORRUPTION
BAD CAL DATA
User calibration corrupt – repeat
calibration
UTILITY DATA
CORRUPTION
BAD UTI DATA
Utility file corrupt – renew all entries
ALARM/RELAY DATA
CORRUPTION
BAD RLY DATA
Alarms/Relays file corrupt-renew all
entries
ANALOGUE ASSIGN
DATA CORRUPTION
BAD ANL DATA
Analogue output file corrupt-renew
all entries
BAD SPAN IGNORED
BAD SPAN
Coarse span adjustment limit
exceeded – Check span sample
NOTE
DETECTOR SIGNAL LOW may briefly occur at power-up.
5.5
5.4
Serious Fault (Shutdown) Conditions
If the Fault LED is illuminated and the Analogue Outputs have been driven HIGH (or
LOW, depending on selection), a Serious Fault exists and the Measurement Display
will show "Measurement Invalid".
Immediate action should be taken to rectify the fault, and qualified service personnel
should be called.
The SERIOUS FAULT messages listed in Table 5.4 will result in a Shutdown.
Table 5.4 – Serious Fault Displays
Display Message
History Log
Entry
Description/comments
INFRARED SOURCE
FAILURE
SOURCE FAIL
Source failure
CHOPPER MOTOR
OUT OF LOCK
OUT OF LOCK
Motor or timing failure
DETECTOR SIGNAL
HIGH
DET SIG HI
Possible loss of IR filter, loss of
liquid sample.
NOTE
•
Detector Signal High may also occur and cause temporary shutdown on
liquid analysers if there is no liquid sample in the cell, e.g. at initial power-up.
Check for loss of flow if sensor fitted. Try Manual Zero on good zero sample if
sample flowing.
•
The user interface remains operational during Shutdown and can be used to
diagnose the likely cause of the fault via the Fault Display and Fault History.
5.5
Cell Maintenance
As a guide the cell should be removed for cleaning when the Reference detector
signals (Section 5.2) fall below 50% of their original values. In aggressive sample
conditions, the loss of signal may be due to general corrosion or pitting of the cell
windows. In this case the cell windows may have to be replaced. Otherwise, the cell
and windows can be dismantled and carefully cleaned and all items (except O-rings)
re-used. See Figure 5.1 for general construction of the cell.
5.6
Figure 5.1
General Cell Construction
CAUTION
Use only Servomex supplied spare parts. The use of inferior replacement
components may degrade the performance and safety of the instrument.
5.7
WARNING
•
Lethal voltages: mains AC power supplies are potentially lethal. The
maintainer must ensure that all mains power supplies are disconnected
from the supply before maintenance work is started.
•
Purge the measuring cell and associated pipework with an inert gas
before work is started.
•
It is the users responsibility to ensure that work is carried out in an
adequately ventilated space – taking due account of risks associated
with release of flammable or toxic samples. Attention is also drawn to
the risks associated with the release of asphyxiants.
To remove the cell assembly from the 2500 proceed as follows:
Flush the cell thoroughly with Nitrogen and disconnect the process connections.
Where fitted, remove cell heater connections:
Figure 5.2
5.8
Suppressor Mounting Details
Unscrew the cell heater wiring connections from inside the source enclosure at terminal
blocks TB32-4 and TB33-4. Loosen the two suppressors from the mounting bracket
and undo the earth terminal (refer to Figure 5.2, taking care not to drop the screw or
the shakeproof washer into the enclosure). Remove the cable ties that hold the ferrite
and slide it off of the wires. Loosen the gland nut on the exterior of the enclosure and
carefully feed the suppressors through the gland one at a time.
Allow sufficient time for the cell assembly to cool (this may take some hours).
Remove any purge connections from the end bosses if fitted. Slacken 3 x M6 socket
screws securing the cell bosses at either end of the sample cell.
Carefully disconnect the cell thermocouple from the cell body. Support the cell
assembly and pull the bosses inward from the ends of the 2500 sufficient to enable the
cell to be removed and taken to a workshop for maintenance. If necessary, slacken the
screws holding the source end casting onto the mounting chassis to permit easy
removal.
CAUTION
Do not pull or strain heater connection leads at any stage.
If the scrubbers (if fitted) are not to be removed, leave the bosses on the analyser and
blank the open ends of the bosses using the 2 blanking plugs (02500460), supplied
with the Special Tools Kit S2500979, to preserve the existing scrubbers. Otherwise
take the entire cell and boss assembly to a workshop.
Working on a clean bench in a suitable environment, dismantle the cell. Pull the end
bosses off the cell body. Note that these items are one-piece stainless steel. If the
bosses are fitted with absorbent scrubbers, these should be removed and replaced
(A variety of types may be fitted – see spares list for appropriate part numbers).
WARNING
Special materials are used for the windows. Some of these are toxic (e.g.
barium fluoride, zinc selenide, etc.) these substances must be handled in
accordance with the relevant local standards.
The windows are usually ground from optical grade single crystal material and as such
may be EXTREMELY EXPENSIVE. The windows may have special anti-reflective
coatings and they may easily be damaged by brittle fracture. The standard material
chosen by Servomex is Calcium Fluoride. This material is non-toxic and reasonably
robust. However, some applications may require materials such as Barium Fluoride,
Zinc Selenide, etc. which are toxic and must be handled with extreme care. See the
Manufacturing Data Sheet at the front of this manual for details of the windows fitted.
Windows should be kept optically clean, i.e. free from dirt, scratches, grime, grease,
fingerprints etc. When handling windows it is recommended that cotton gloves or finger
cots be worn.
5.9
WARNING – ZINC SELENIDE
If windows of zinc selenide are fitted and have been damaged by acidic attack
(e.g. acidic condensate or organic acids have entered the cell) it will be likely
that some of the zinc selenide will be reduced to selenium metal. This is
highly toxic and on no account should attacked zinc selenide material be
handled with bare fingers. Such material should be bagged and disposed of
according to local regulations.
Cleaning of the cell windows should be carried out using cotton buds or lens tissue,
detergent and/or a suitable solvent such as Isopropanol. Avoid abrasive cleaning
compounds, as these will damage the optical surface.
WARNING
The use of some solvents (iso-propyl alcohol, alcohol etc.) constitutes a
hazard. These substances must be handled in accordance with the relevant
local standards.
To remove the cell windows proceed as follows:
Ensure that the end bosses have been removed from the cell body. Starting at one end
first, unscrew and remove one locking ring with a cell key (supplied in Special Tools Kit
Servomex p/n S2500979). Remove the 5 wavy washers, spacer, and 'O' ring being
careful not to damage them or the window. To remove the window, seal the outlet and
with the cell in a vertical position, connect a hand operated aspirator (Servomex p/n
2387-0514) to the inlet. Apply a gentle pressure to lift the window up squarely and
remove it. Finally remove the PTFE gasket. Dismantle the other end of the cell in a
similar manner.
The 'O' ring spacer is threaded internally. The Special Tools Kit Servomex p/n
S2500979 contains an extractor, which screws into the spacer to aid removal.
The windows may now be carefully cleaned. The cell body may also be cleaned as
necessary.
Ensure all items are thoroughly clean and dry before reassembling, which is the
reverse procedure but with the following points noted:
•
The windows should be 'floated' into place squarely using a hand aspirator or
equivalent, reversing the removal procedure.
•
Use new 'O' rings and other cell sealing kit items making sure that all dust, hairs
etc, are rigorously excluded from them. Ensure the window surfaces are also
free from solvent residues or finger grease.
•
The locking ring has a coated thread to prevent seizure – PTFE tape is not
required. The locking ring should be tightened down in stages, pausing for a
moment between stages until the metal faces of the cell body and locking ring
meet when fully tightened, with no gap between them. Where Teflon 'O' rings
5.10
are being used, this will require significant torque, with longer pauses. DO NOT
use the cell inlet/outlet stubs as levers.
After re-assembly, perform a leak check on the cell using a manometer before installing
it in the 2500. Do not submerge the cell or fill it with water.
Re-installation of the complete cell assembly onto the 2500 is the reverse of the
dismantling process with the added requirement that, unless fitted with a purge, the end
bosses should be briefly flushed with clean dry nitrogen after the cell has been
mounted and then new absorbent scrubbers should be quickly screwed into them. Be
sure to fully tighten the screws holding the source end casting onto the mounting
chassis if they have been slackened.
Finally, connect up the cell thermocouple to the cell body, reconnect the cell heater
connections (where fitted) remembering to refit the ferrite and secure it with cable ties
as close as possible to the suppressor bracket (see Figure 5.2 for details). Connect up
any purge connections and the process connections. Return the analyser to service as
described in Section 4.5.
Reset zero if necessary on a true zero sample.
5.6
Routine Leak Checks
WARNING
It is the users responsibility to keep all sample connections to the 2500
sample cell, purge connections (if fitted) to the end bosses and steam coil
connections completely free from leaks. All plumbing connections should
therefore be regularly inspected for signs of leakage and urgent corrective
action taken if any are found. This is especially important for process
samples, which are flammable, toxic, liquid, or any combination of these.
5.11
5.12
SECTION 6
6.1
SPARES LISTS
Spares List
The following spares list applies to all versions of the 2500 covered by this manual.
The 2500's Model and Serial Number MUST be advised when ordering spares.
NOTE
For certified versions EU1, EU2, DU1, EX1, EX2 and DX1, only the boards listed
may be used as spares
6.1.1
General
Table 6.1 – General Spares List
Description
Part Number
Power Supply
S2000925
Display PCB
02501903
Display/Keypad Assembly
S2501999
Keypad Assembly
02501998
Sensor Interface PCB
02000934
Option PCB
02000916
Transmitter PCB
S2500904A
Motor Connection PCB
S2500906
Power Control PCB
S2501911A
Interconnecting Ribbon Cable Assembly 1-256mm path
S2500928A
Interconnecting Ribbon Cable Assembly 512mm path
S2500928B
Interconnecting Ribbon Cable Assembly 1000mm path
S2500928C
Display Ribbon Cable Assembly
S2500929
Interconnecting Cableform 1-256mm path
S2500944A
Interconnecting Cableform 512mm path
S2500944B
Interconnecting Cableform 1000mm path
S2500944C
Fixings Kit (All fasteners)
S2500974
Seals Kit (All Seals excluding Cell Seals)
S2500975
Chopper Box Insulation
S2500977
Transformer, 115/230 volts
00608333
6.1
Table 6.1 – General Spares List
6.1.2
Transformer, 100/200 volts
00608334
Special Tools Kit
S2500979
Enclosure Earth Studs
S2500984
Pressure Transducer
S2500987
Chopper Motor
S2500986
Chopper Box Seals Kit
S2500988
Gas Strut Kit
S2500989
Fuse Kit
02501996
Multi component Adaptor PCB (2550 only)
S2550901
10 x Gas Filter Correlation Adaptor PCB (2510 only)
S2510901A
40 x Gas Filter Correlation Adaptor PCB (2510 only)
S2510901B
Processor PCB + Firmware (2510 only)
S2510965B
Processor PCB + Firmware (2500, 2520, 2550)
S2500965B
Thermal Fuse (98°C)
2536-0329
Communications Board (RS485)
S2500912
Communications Board (Ethernet)
S2500913
Sample Cell Spares
Table 6.2 – Sample Cell Spares
6.2
Description
Part Number
Thermocouple (300mm Sheath)
02500502A
Cell Servicing Kit
S2500981
Cell Seals Kit (Viton + PTFE)
S2500982A
Cell Seals Kit (Chemraz)
S2500982B
Special Tools Kit
S2500979
PTFE Tape
1835-3026
Hand Aspirator
2387-0514
6.1.3
Source Units
(Refer to Final Specification Sheet)
Table 6.3 – Source Unit Spares
6.1.4
Description
Part Number
Long Wave, CaF2
S2500931C
Long Wave, BaF2
S2500931B
Long Wave, ZnSe
S2500931A
Short Wave, CaF2
S2500943A
Short Wave, Glass
S2500943B
Soft UV CaF2 (2520 Only)
S2520933A
Detector Units
(Refer to Final Specification Sheet)
Table 6.4 – Detector Spares
Description
Part Number
CaF2
S2501905A
CVDFET
S2501905C
KRS-5
S2501905B
Photodiode (2520 Only)
S2600905
6.3
6.1.5
Windows
(Refer to Final Specification Sheet)
Table 6.5 – Window Spares
6.1.6
Description
Part Number
CaF2
S2500980A
BaF2
S2500980B
ZnSe
S2500980C
Quartz
S2500980D
Glass
S2500980E
Sapphire
S2500980G
Scrubbers
(Refer to Final Specification Sheet)
Table 6.6 – Scrubber Spares
6.4
Detector / Cell Boss Scrubber (H2O) Marked 'W'
S1200921
Detector / Cell Boss Scrubber (H2O/CO2) Marked 'CW'
S1200922
Detector / Cell Boss Scrubber (H2O/CO2/CO) Marked 'H'
S1200924
Scrubber Sachet Kit, H2O Mix
1723-8010
Scrubber Sachet Kit, H2O/CO2 Mix
S2000511
Scrubber Sachet Kit, H2O/CO2/CO Mix
S2000512
6.2
Recommended Spares
Quantity for 2 years.
Table 6.7 – Recommended Spares
Item
Part Number
1-3 Analysers
4-9 Analysers
Fuse Kit
02501996
2
4
Cell Servicing Kit
S2500981
2
4
PTFE Tape
1835-3026
1
1
Chopper Motor
S2500986
1
1
Cell sealing Kit
Refer to application
2
5
Hand Aspirator
2387-0514
1
1
Windows
Refer to application
2
4
Chopper Box Scrubber
Refer to application
2
4
Detector & Cell Boss
Scrubber
Refer to application
2
4
Source
Refer to application
1
2
Detector
Refer to application
0
1
Microprocessor PCB
Refer to application
0
1
Sensor Interface PCB
02000934
0
1
Power Supply
S2000925
0
1
NOTE
Scrubbers should be used as soon as received, or otherwise they must be stored in
their original unopened packaging in an airtight container, or preferably in a
desiccator. Scrubbers should be used on a "first in, first out" basis.
6.5
6.6
SECTION 7
INSTRUMENT SPECIFICATIONS
7.1
Generic 2500 Series Performance
7.1.1
Environmental Specifications
Table 7.1 – General Environmental Specifications
Operating Temperature
GEN, EU1, EU2, DU1, EX1, EX2, DX1,
HTV 0-55°C (32-131°F)
Operating Humidity
0-95% RH, non-condensing
Storage Temperature
-25°C(-13°F) to +70°C (158°F)
Storage Humidity
0-95% RH, non-condensing
Altitude
Up to 3000 M
Installation Category II
Pollution Degree II
Over-voltage impulse to withstand up to 2500V
in accordance with IEC 60664-1
Rate of Ambient Temperature
Change
Less than ±25°C/h, (±45°F/h) Change
Warm-up Time
Typically 2-10h, depending on application and
environment
7.1.2
Dimensions
Length:
Max 1618mm, (63.7") min 615mm (24.2")
Height:
241mm (9.5")(allow 500 mm (20")
to open end covers)
Width:
248mm (9.8")
Weight:
from 25 kg (55 lb) to 50 kg (110 lb)
7.1.3
Power Supply
115/230V AC ±15%, 50/60Hz or 100/200V AC ±15%, 50/60Hz.
120 VA without optional electrically heated cell.
300 VA with optional electrically heated cell.
Power Supply
Voltage Effects:
Less than ±1% fsd for ±15% change.
Power Supply
Frequency Effects:
Less than ± 1% fsd for 47 to 53 Hz or 57 to 63 Hz.
7.1
7.1.4
Performance Characteristics (each component)
Table 7.2 – Performance Characteristics
Intrinsic Error*
Less than ±1% fsd
Repeatability
±0.5% fsd
Output Fluctuation (Noise)
Less than 1% fsd (peak-peak) at minimum T90†
Linearity Error
Typically less than ±1% fsd
Short Term Zero Drift:
Less than ±1% fsd per week (2500)
Less than ±1% fsd per week (2510 - for ranges greater than
100ppm fsd)
Less than ±2% fsd per week (2520, 2550)
Less than ±2% fsd per week (2510 - for ranges less than or
equal to 100ppm fsd)
Response Time (T90)
User adjustable from 11s minimum (electronic
only, excludes sampling)
Obscuration
Less than ±3% fsd zero error for 50% broadband
obscuration of cell windows
Influence Errors
Dependant on application (see specification
details)
* error when used under reference conditions.
†
2510 HCL measurement less than 2% fsd (peak-peak) at minimum T90.
7.1.5
Performance Characteristics – EMC
Configuration of equipment:
Fitted with an unscreened power cable. Also included were two separate mA
output cables, each individually screened. Each signal pair was fitted with
ferrite bead (Servomex part No. 2824-0017). The screens and drain wires were
terminated at the glands.
Results:
Met the requirements of EN 61326, Table 4, Class B (Equipment for use in
domestic establishments) in respect of conducted and radiated emissions.
Met the requirements of EN 61326, Table A1 (Equipment for use in industrial
establishments) in respect of electrostatic discharges, fast transients/bursts,
surges, conducted RF disturbances, voltage dips and interrupts.
An additional error of 4% FS over the intrinsic error given in Table 7.2 may be
observed at some frequencies under the influence of radiated RF fields
specified for industrial environments.
Connection of additional or other signal/control cables and pressure
compensation could lead to a deterioration in performance at some RF
frequencies.
7.2
7.1.6
Sample Specification
Table 7.3 – Sampling Specification
Sample Temperature
0-180°C/32-356°F (see sampling influences)
Sample Pressure
0-150 psig/0-10 barg (normal) (For special high
pressure operation consult Servomex)
Sample Flow (typical)
GAS 0.2-5.0 litres/min
LIQUIDS 0.3-1.0 litres/min
(Constant flow rate preferred)
Sample Wetted Materials
Refer to Final Specification Sheet
7.1.7
Sample and Ambient Performance
Table 7.4 – Sample and Ambient Influence
Sample Pressure
Stability
Output may change up to ±1.6% fsd for a ±1% change in
sample (Gas Analysers) pressure. With optional
empirical sample pressure compensation, effect is
reduced to ±1% fsd or less for a 20% change in sample
pressure for most typical samples.
Sample Temperature
Stability
Output may change up to ±0.3% fsd for gases for a ±1°C
change in sample temperature. With optional empirical
sample temperature compensation, effect is reduced to
±1% fsd or less for a 20°C change in sample
temperature for most typical samples.
Sample Flow Stability
Depends on cell volume and vent conditions but for
unheated cells the primary effect is less than ±1% fsd
change in output for flow rate changes of ±10%.
Ambient Temperature
Influence (2500 & 2550)
Less than 1% fsd zero drift due to rate of ambient
temperature change of 25°C/hr (45°F/hr) over a
maximum of 25°C (45°F) change.
Ambient Temperature
Influence (2510)
3% fsd per 10°C (18°F) change.
Ambient Temperature
Influence (2520)
Less than 2% fsd zero drift due to rate of ambient
temperature change of 25°C/hr (45°F/hr) over a
maximum of 25°C (45°F) change.
7.1.8
Solenoid Valve Relays
3 x Solenoid valve relays with volt free contacts rated 240VAC/1.0A, 30VDC/1.0A.
1 x 24VDC/12VA auxiliary supply for solenoid valves
7.3
7.1.9
Analogue Outputs
2 x mA outputs, plus 3 x relay outputs. Optionally additional 2 x mA, plus 2 x relay
outputs.
All analogue mA outputs are assignable and selected as 0-20mA or 4-20mA (max.
impedance 1KOhm) or alternatively 0-10V or 2-10V (min. impedance 1MOhm).
Dual ranges on every Analogue output. Optional additional mA outputs should be
specified for 3 component analysers.
All relay outputs are assignable contact pairs selected by links as NC or NO, and are
rated 240VAC/1.0A, 30VDC/1.0A.
7.1.10
Serial Output
Single ASCII data logging output port (RS232).
Table 7.5 – Serial Output Specification
7.1.11
Baud Rate
2400
Parity
Even
Stop Bits
1
Data Bits
8
Handshake
Hardware
Modbus RS-485
Modbus serial communication over RS-485 interface using option board. Note that
these settings are independent from those in section 7.1.10
Table 7.6 – Modbus RS-485 Specification
Protocol
Modbus ASCII & RTU
Modbus Address
1 to 247
Default = 247
Baud Rate
9K6/4K8/2K4/19K2
Default = 9K6
Parity
EVEN/ODD/NONE
Default = EVEN
Stop Bits
1/1.5/2*
Default = 1
Data Bits
7/8*
Default = 7 (ASCII), 8
(RTU)
* Note:- Stop & Data bits may only be set if Modbus ASCII mode is selected. Modbus
RTU will use fixed settings of Data = 8 and Stop = 2 (for parity = NONE) or Stop = 1
(for parity not set to NONE).
7.4
7.1.12
Ethernet
Communication over Ethernet using option board.
This provides a 10/100M Ethernet interface and uses the Modbus TCP protocol.
User interface allows the IP address, subnet mask and gateway address to be entered.
All fields default to zeros. A provisional subnet mask is generated whenever an IP
address is entered that falls into a different class. This can then be modified if required.
7.1.13
Inputs
Terminals for external switch contacts are provided for:
•
Range change
•
Autocalibration/Autocheck initiate
•
Key switch (password)
•
Flow Sensor
7.1.14
Ingress Protection
Designed to IP65 (without electrical cell heater, IP50 with electrical cell heater).
7.1.15
Optical Purge
Where used, end bosses require total 200cm3/min clean, dry purge gas (see Section 4.2)
delivered at 1-3 psig.
7.1.16
Alternative Purge Controller (2500 EU1 & EX1 Only)
Minimum flow rate = 10 SLPM for 11 minutes.
Minimum overpressure = 0.5 mBar.
Maximum overpressure = 80 mBar.
Refer to Certification Manual (Certificate SIRA 09ATEX1341X and IECEx
SIR09.0136X) for definitive requirements and operational parameters.
7.5
7.6
SECTION 8 CE MARKING AND OTHER SAFETY APPROVALS
8.1
•
•
EMC and Electrical Safety
The Analyser complies with the European Community “Electromagnetic
Compatibility Directive”:
•
Emissions: Equipment suitable for use in domestic establishments and in
establishments directly connected to a low voltage supply which supplies
buildings for domestic purposes.
•
Immunity: Industrial locations.
The analyser complies with the European Community “Low Voltage Directive”
by the application of:
•
EN61010-1 and rated for Over-Voltage Category II, Pollution Degree 2.
•
The analyser complies with the class B digital apparatus requirements of ICES003 of Canada through the application of EN 55022:1994.
•
L’analyseur est conforme aux Conditions B numériques d’appareillage de
classe de NMB-003 du Canada par l’application du EN 55022:1994.
•
The analyser complies with part 15 of the US FCC Rules for Class B
equipment. It is suitable for operation when connected to a public utility power
supply that also supplies residential environments.
8.1
8.2
ATEX Directive and other non-European Hazardous Area approvals
ATEX Directive 94/9/EEC
NOTE
Depending on the variant ordered an additional Safety Certificate Manual may be
supplied. This contains all the relevant safety certificates and declarations for the
EU1, EU2, EX1 and EX2 variants.
WARNING
The installer and/or the user of the analyser must be satisfied that the
installation complies with any "Special Conditions for Safe Use" or
"Schedules of Limitations" contained in the Certificates included in the
separate certification manual.
Table 8.1 – 02500 Safety Approvals
Variant
GEN/HTV
EU1
Certification
Only suitable for installation in a non-hazardous area.
The 2500 Series EU1 version is ATEX CAT 2 equipment
certified to Type 'p' standard EN 60079-2 and is suitable
for use in European areas that require EPL Gb. The
analyser is marked Ex px ia [ia] IIC T5 Gb for unheated
cells and Ex px ia [ia] IIC T4 Gb for steam heated cells.
The 02500 Series EU2 version is ATEX CAT 3
equipment certified to Type 'n' standard EN 60079-15,
and is suitable for use in areas that require EPL Gb. The
analyser is marked Ex nCL IIC T3 Gc for cell
temperatures up to 130°C and Ex nCL IIC T4 Gc for cell
temperatures up to 80°C.
EU2/DU1
The 02500 Series EU2 version is ATEX CAT 2
equipment certified to Dust standard IEC 60079-31 and is
suitable for use in areas that require ELP Db. The
analyser is marked Ex tb IIIC T175°C Db IP66 for cell
temperatures up to 130°C and Ex tb IIIC T125°C Db IP66
for cell temperatures up to 80°C and Ex tb IIIC T80°C Db
IP66 for unheated cells.
Special condition of this certificate is:
To avoid propagating brush discharges, the unit is not to
be installed in a high velocity dust-laden atmosphere.
8.2
Table 8.1 – 02500 Safety Approvals
EX1
The 02500 Series EX1 version is IECEx certified to Type
‘p’ standard IEC60079-2 and is suitable for use in areas
that require EPL Gb. The analyser is marked Ex px ia
[ia] IIC T5 Gb.
The 02500 Series EX2 version is IECEx certified to Type
‘n’ standard IEC 60079-15 and is suitable for use in areas
that require EPL Gc. The analyser is marked Ex nCL IIC
T3 Gc for cell temperatures up to 130oC and Ex nCL IIC
T4 Gc for cell temperatures up to 80oC or unheated cells.
EX2/DX1
The 02500 Series EX2 version is also IECEx certified to
Dust standard IEC 60079-31 and is suitable for use in
areas that require EPL Db. The analyser is marked Ex tb
IIIC T175oC Db IP66 for cell temperature up to 130oC
and Ex tb IIIB T125oC Db IP66 for cell temperatures up
to 80oC and Ex tb IIIB T80oC Db IP66 for unheated cells.
Special condition of this certificate is:
To avoid propagating brush discharges, the unit is not to
be installed in a high velocity dust-laden atmosphere.
8.3
8.4
APPENDIX A- Modbus Data Mapping
When fitted with one of the optional communication boards, the 2500 series analysers
are capable of bi-directional digital communications based on the Modbus protocol.
The following facilities are provided:•
Access to measurements
•
Access to measurement diagnostic values
•
Access to alarm states
•
Access to detailed fault flags
•
Ability to start and stop autocalibration
•
Ability to read or modify autocalibration setup
•
Ability to monitor autocalibration progress
Access to these facilities is achieved through reading (and in some cases writing)
Modbus registers and coils / discrete inputs as detailed below.
A.1
Floating Point Numbers
All floating point numbers are represented in 32 bit IEEE754 format and must therefore
be transmitted in Modbus messages as two registers.
The following coil / discrete input controls the order in which pairs of registers
representing floating point values is transmitted. This control setting applies to all
floating point values read and written.
This bit may be set using function code 05. Its current state may be read with function
codes 01 and 02.
Coil/
Discrete I/P
Number
Coil/
Discrete I/P
Address
Name
301
300
Floating Point
Register Order
Comments
Changes the order of the Modbus register when
dealing with 32-bit floating point numbers.
0= Big-endian, e.g.
Register number 1 = high word
Register number 2 = low word (default)
1 = Little-endian, e.g.
Register number 1 = low word
Register number 2 = high word
A.1
A.2
Measurement Summary
This block of registers allows all primary measurement components to be read in single
query.
This register data is read-only and may be accessed with function codes 03 and 04.
Register
Number
Register
Address
Name
Data
Type
Comments
1
0
Number of
measurements
integer
2-3
1-2
Measurement 1
float
Measurement component 1
4-5
3-4
Measurement 2 *
float
Measurement component 1
6-7
5-6
Measurement 3 *
float
Measurement component 1
Number of fitted measurement
components
* Note that measurements 2 and 3 are optional and may only be present in 2550
instruments. Where not fitted these values will be zero.
A.3
Measurements Detail
This register data is read-only and may be accessed with function codes 03 and 04.
Register
Number
Register
Address
101-102
100-101
103-105
Name
Data
Type
Comments
Measurements 1
float
Measurement 1 value
102-104
Measurement 1 formula
string
Measurement 1 chemical formula
(max 6 chars)
106-107
105-106
Measurement 1 units
string
Measurement 1 units (max 3 chars)
108-110
107-109
Reserved
111-112
110-111
Measurement 2 *
float
Measurement 2 value
113-115
112-114
Measurement 2 formula*
string
Measurement 2 chemical formula
(max 6 chars)
116-117
115-116
Measurement 2 units *
string
Measurement 2 units (max 3 chars)
118-120
117-119
Reserved
121-122
120-121
Measurement 3 *
float
Measurement 3 value
123-125
122-124
Measurement 3 formula*
string
Measurement 3 chemical formula
(max 6 chars)
126-127
125-126
Measurement 3 units *
string
Measurement 3 units (max 3 chars)
128-130
127-129
Reserved
* Note that measurements 2 and 3 are optional and may only be present in 2550
instruments. Where not fitted these values will be zero.
A.2
A.4
Measurement Diagnostics
This register data is read-only and may be accessed with function code 03 and 04.
Register
Number
Register
Address
201-202
200-201
Sample pressure
float
Sample pressure (zero if not fitted)
203-204
202-203
Chopper temperature
float
Chopper temperature as oC
205-206
204-205
Compensation temperature
float
Compensation (detector) temperature as oC
207-208
206-207
Cell/Sample temperature
float
Cell/Sample temperature as oC
209-210
208-209
Source voltage
float
Source voltage
211-212
210-211
Intensity 1
float
213-214
212-213
Intensity 2
float
215-216
214-215
Intensity 3
float
217-218
216-217
Intensity 4
float
219-220
218-219
Intensity 5
float
221-222
220-221
Intensity 6
float
223-230
222-229
Reserved
A.5
Name
Data
Type
Comments
Intensity channel values representing the input
positions used for measurement and reference
signals.
Alarms
This coil / discrete input data is read-only and may be accessed with function code 01
and 02.
Coil/Discrete
I/P Number
Coil/Discrete
I/P Address
Name
1
0
Measurements 1 Alarm 1
2
1
Measurements 1 Alarm 2
3
2
Measurements 1 Alarm 3
4
3
Measurements 1 Alarm 4
5-8
4-7
9
8
Measurements 2 Alarm 1 *
10
9
Measurements 2 Alarm 2 *
11
10
Measurements 2 Alarm 3 *
12
11
Measurements 2 Alarm 4 *
13-16
12-15
17
16
Measurements 3 Alarm 1 *
18
17
Measurements 3 Alarm 2 *
19
18
Measurements 3 Alarm 3 *
20
19
Measurements 3 Alarm 4 *
21-24
20-23
Reserved
Reserved
Reserved
Comments
Alarm condition indicated by bit = 1
All bits = 0
Alarm condition indicated by bit = 1
All bits = 0
Alarm condition indicated by bit = 1
All bits = 0
A.3
* Note that measurements 2 and 3 are optional and may only be present in 2550
instruments. Where not fitted these values will be zero.
A.6
Fault Data
This coil / discrete input data is read-only and may be accessed with function code 01
and 02.
Coil/Discrete
I/P Number
Coil/Discrete
I/P Address
101
100
Fault status
102
101
Optical bench power failure
103
102
Source voltage high
104
103
Source voltage low
105
104
Infrared source failure
106
105
Chopper motor out of lock
107
106
Sample flow failure
108
107
Chopper temperature high
109
108
Chopper temperature low
110
109
Chopper temperature sensor
failure
111
110
Cell temperature high
112
111
Cell temperature low
113
112
Cell temperature sensor
failure
114
113
Sample temperature sensor
failure
115
114
Pressure sensor failure
116
115
Bad reference voltage
117
116
Compensation temperature
sensor failure
118
117
Detector signal high
119
118
Detector signal low
120-122
119-121
123
122
Auto-cal bad pre flush
124
123
Zero out of tolerance
125
124
Span out of tolerance
126-130
125-129
A.4
Name
Reserved
Reserved
Comments
Bit = 1 indicates fault condition present
Bit = 1 indicates specified fault condition
present
All bits = 0
Bit = 1 indicates specified fault condition
present
All bits = 0
131
130
Password violation
132
131
System clock inoperative
133
132
System ADC inoperative
134
133
Default system data
corruption
135
134
Calibration data corruption
136
135
Utility data corruption
137
136
Alarm / relay data corruption
138
137
Analogue assignment data
corruption
139
138
Bad span ignored
A.7
Bit = 1 indicates specified fault condition
present
Autocalibration Setup
These registers allow the autocalibration settings to be read using function codes 03
and 04 and to be written using function codes 06 and 16 (see note 1).
Register
Number
Register
Address
Name
Data
Type
Comments
301
300
Select low / high
Integer
Zero = low autocalibration only.
Non-zero = low & high calibration.
302
301
Mode
Integer
Zero = check only.
Non-zero = calibrate
303-304
302-303
DV lag
float
Required delay in minutes
305-310
304-309
Reserved
311-312
310-311
Measurement 1 low target
float
Concentration of low calibration gas for
measurement 1
313-314
312-313
Measurement 1 high target
float
Concentration of high calibration gas for
measurement 1
315-320
314-319
Reserved
321-322
320-321
Measurement 2 low target
(see note 2)
float
Concentration of low calibration gas for
measurement 2
323-324
322-323
Measurement 2 high target
(see note 2)
float
Concentration of high calibration gas for
measurement 2
325-330
324-329
Reserved
331-332
330-331
Measurement 3 low target
(see note 2)
float
Concentration of low calibration gas for
measurement 3
333-334
332-333
Measurement 3 high target
(see note 2)
float
Concentration of high calibration gas for
measurement 3
A.5
Notes:1.
Any attemp to write a register in the above range, whilst an autocalibration is in
progress, will return an exception response 06 (slave busy). Register values will
not be updated.
2.
Measurements 2 & 3 are optional and may only be present in 2550 instruments.
Where not fitted reading these values will return zero.
3.
The DV lag should be expressed in units of minutes to the nearest 0.5 minutes.
Other values will be automatically rounded to conform before being written.
A.8
Autocalibration control
The following coils / discrete inputs control the autocalibration process. Control actions
are initiated by setting the specified bit = 1 for between 2 and 59 seconds and then
setting the bit = 0. The autocalibration configuration, including the selection of
calibration or check, is determined by the analyser settings.
These bits may be set using the function code 05. Their current state may be read with
function codes 01 and 02.
Coil/Discrete
I/P Number
Coil/Discrete
I/P Address
201
200
Start autocalibration
Initiates an autocalibration/autocheck using the
predefined calibration setting.
202
201
Stop autocalibration
Terminate a running autocalibration / autocheck
process (however initiated).
A.6
Name
Comments
A.9
Autocalibration Status
This register data is read-only and may be accessed with function code 03 and 04.
Register
Number
Register
Address
401
400
402-403
401-402
Name
Autocalibration Phase
Data
Type
Comments
Integer
Value indicates autocalibration phase as
follows:0 = Not in autocalibration
1 = In span pre flush
2 = In zero calibration
3 = Zero corrected
4 = In span calibration
5 = Span corrected
6 = In post-flush
Measurement 1 zero precalibration
float
404-405
403-404
Measurement 1 zero postcalibration
float
406-407
405-406
Measurement 1 Span precalibration
float
Measurement 1 Span
post-calibration
float
408-409
407-408
410
409
Reserved
float
411
410
Autocalibration phase *
float
412-413
411-412
Measurement 2 zero precalibration *
float
414-415
413-414
Measurement 2 zero postcalibration *
float
416-417
415-416
Measurement 2 Span precalibration *
float
Measurement 2 Span
post-calibration *
float
418-419
417-418
420
419
Reserved
421
420
Autocalibration Phase *
Integer
These values are updated during phase
3 of the autocalibration sequence.
These values are updated during phase
5 of the autocalibration sequence.
Value indicates autocalibration phase as
follows:0 = Not in autocalibration
1 = In span pre flush
2 = In zero calibration
3 = Zero corrected
4 = In span calibration
5 = Span corrected
6 = In post-flush
These values are updated during phase
3 of the autocalibration sequence.
These values are updated during phase
5 of the autocalibration sequence.
Value indicates autocalibration phase as
follows:0 = Not in autocalibration
1 = In span pre flush
2 = In zero calibration
3 = Zero corrected
4 = In span calibration
5 = Span corrected
6 = In post-flush
A.7
422-423
421-422
Measurement 3 zero precalibration *
float
424-425
423-424
Measurement 3 zero postcalibration *
float
426-427
425-426
Measurement 3 Span precalibration *
float
Measurement 2 Span
post-calibration *
float
428-429
427-428
These values are updated during phase
3 of the autocalibration sequence.
These values are updated during phase
5 of the autocalibration sequence.
* Note that measurements 2 and 3 are optional and may only be present in 2550
instruments. Where not fitted these values will be zero.
A.8
APPENDIX B - Equipment Protection Levels.
Traditional Relationship of Equipment Protection Levels (EPLs) to Zones
Equipment Protection Level (EPL)
Zone
Ga
0
Gb
1
Gc
2
Da
20
Db
21
Dc
22
For more information on Equipment Protection Levels (EPLs) refer to Annex D of IEC
60079-0:2007 or EN60079-0:2009.
B.1
B.2
Appendix C - Disposal In Accordance With The Waste Electrical and Electronic
Equipment (WEEE) Directive.
The analyser is not considered to be within the scope of the Waste Electrical and
Electronic Equipment (WEEE) Directive.
The analyser is not intended for disposal in a municipal waste stream, but shall be
submitted for material recovery and recycling in accordance with any appropriate local
regulations.
For additional information and advice on the disposal of the analyser, contact
Servomex.
Servomex Group Limited
Jarvis Brook
Crowbrough
East Sussex
TN63FB
England
Tel: (44) 1892 652181
Fax: (44) 1892 662253
Global email: info@servomex.com
If you send the analyser to Servomex or your local Servomex analyser for disposal, it
must be accompanied by a correctly completed decontamination certificate.
C.1