Endress + Hauser Operating Instructions Turbimax CUS52D Turbidity sensor
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
BA01275C/07/EN/01.13
71237021
Products
Solutions
Operating Instructions
Turbimax CUS52D
Turbidity sensor
Services
Document information
Warnings
The structure, signal words and safety colors of the signs comply with the specifications of
ANSI Z535.6 ("Product safety information in product manuals, instructions and other
collateral materials").
Safety message structure
!
DANGER
Causes (/consequences)
Consequences if safety
message is not heeded
‣ Corrective action
!
WARNING
Causes (/consequences)
Consequences if safety
message is not heeded
‣ Corrective action
!
CAUTION
Causes (/consequences)
Consequences if safety
message is not heeded
‣ corrective action
NOTICE
Cause/situation
Consequences if safety
message is not heeded
‣ Action/note
Meaning
This symbol alerts you to a dangerous situation.
Failure to avoid the situation will result in a fatal or serious
injury.
This symbol alerts you to a dangerous situation.
Failure to avoid the situation can result in a fatal or serious
injury.
This symbol alerts you to a dangerous situation.
Failure to avoid this situation can result in minor or more
serious injuries.
This symbol alerts you to situations that can result in
damage to property and equipment.
Icons
Additional information, tips
Permitted or recommended
Forbidden or not recommended
Endress+Hauser
Turbimax CUS52D
Table of contents
1
Basic safety instructions . . . . . . . . . . . . . 4
9
Accessories . . . . . . . . . . . . . . . . . . . . . . . 35
1.1
1.2
1.3
1.4
1.5
Requirements for personnel . . . . . . . . . . . . . . . . . . 4
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Occupational safety . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Product safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2
Incoming acceptance and product
identification . . . . . . . . . . . . . . . . . . . . . . 6
9.1
9.2
9.3
9.4
9.5
9.6
9.7
9.8
Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Mounting material . . . . . . . . . . . . . . . . . . . . . . . . . 35
Holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Compressed air cleaning . . . . . . . . . . . . . . . . . . . . 36
Ultrasonic cleaning . . . . . . . . . . . . . . . . . . . . . . . . . 36
Bubble trap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Calibration set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
2.1
2.2
2.3
2.4
Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . 6
Product identification . . . . . . . . . . . . . . . . . . . . . . . . 6
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Certificates and approvals . . . . . . . . . . . . . . . . . . . . 6
10
Technical data . . . . . . . . . . . . . . . . . . . . 38
3
Installation . . . . . . . . . . . . . . . . . . . . . . . . 7
10.1
10.2
10.3
10.4
10.5
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Performance characteristics . . . . . . . . . . . . . . . . . 39
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Mechanical construction . . . . . . . . . . . . . . . . . . . . 40
3.1
3.2
3.3
3.4
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Installation instructions . . . . . . . . . . . . . . . . . . . . . . 9
Installation examples . . . . . . . . . . . . . . . . . . . . . . 12
Post-installation check . . . . . . . . . . . . . . . . . . . . . 18
4
Electrical connection . . . . . . . . . . . . . . . 19
4.1
4.2
Connecting to the transmitter . . . . . . . . . . . . . . . 19
Post-connection check . . . . . . . . . . . . . . . . . . . . . 19
5
Device description . . . . . . . . . . . . . . . . . 20
5.1
5.2
5.3
5.4
5.5
5.6
5.7
Sensor design . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measuring principle . . . . . . . . . . . . . . . . . . . . . . .
Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cyclic cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Signal filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Calkit CUS52 solid state reference . . . . . . . . . . .
6
Diagnostics and troubleshooting. . . . . 32
7
Maintenance . . . . . . . . . . . . . . . . . . . . . . 33
7.1
Cleaning the sensor . . . . . . . . . . . . . . . . . . . . . . . 33
8
Repair. . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
8.1
8.2
8.3
Spare parts kits . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Endress+Hauser
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
20
20
21
22
30
30
31
3
Basic safety instructions
Turbimax CUS52D
1
Basic safety instructions
1.1
Requirements for personnel
‣ Installation, commissioning, operation and maintenance of the measuring system must
‣
‣
‣
‣
only be carried out by specially trained technical personnel.
The technical personnel must be authorized for the specified activities by the system
operator.
Electrical connection must only be carried out by a certified electrician.
The technical personnel must have read and understood these Operating Instructions
and must follow the instructions they contain.
Faults at the measuring point may only be rectified by authorized and specially trained
personnel.
Repairs not described in the enclosed Operating Instructions may only be carried out at
the manufacturer’s premises or by the Service Organization.
1.2
Designated use
CUS52D is a sensor designed to measure turbidity in drinking water and process water
applications.
The sensor is particularly suited for use in the following applications:
• Final turbidity measurement in the outlet of waterworks
• Turbidity measurement in the inlet of waterworks
• Turbidity measurement at all stages of the process
• Turbidity measurement for filter monitoring and filter backwashing
• Turbidity measurement in drinking water networks
Any use other than that described here compromises the safety of persons and the entire
measuring system and is, therefore, not permitted.
The manufacturer is not liable for damage resulting from improper or non-designated use.
1.3
Occupational safety
As the user, you are responsible for complying with the following safety conditions:
• Explosion protection guidelines (only devices approved for use in explosion hazardous
areas)
• Installation instructions
• Local prevailing standards and regulations
Electromagnetic compatibility
This device has been tested for electromagnetic compatibility in accordance with the
applicable European standards for industrial applications.
The electromagnetic compatibility indicated only applies to a device that has been connected
in accordance with the instructions in these Operating Instructions.
1.4
Operational safety
‣ Before commissioning the entire measuring point, make sure all the connections are
‣
‣
4
correct. Ensure that electrical cables and hose connections are not damaged.
Do not operate damaged products, and secure them against unintentional
commissioning. Mark the damaged product as defective.
If faults cannot be rectified, the products must be taken out of service and secured against
unintentional commissioning.
Endress+Hauser
Turbimax CUS52D
Basic safety instructions
1.5
Product safety
The product is designed to meet state-of-the-art safety requirements, has been tested and
left the factory in a condition in which it is safe to operate.
Relevant regulations and European standards have been observed.
Endress+Hauser
5
Incoming acceptance and product identification
2
Incoming acceptance and product
identification
2.1
Incoming acceptance
Turbimax CUS52D
• Make sure the packaging is not damaged.
• Notify the supplier of any damage to the packaging. Keep the damaged packaging until the
matter has been settled.
• Make sure the contents are not damaged.
• Notify the supplier of any damage to the delivery contents. Keep the damaged products
until the matter has been settled.
• Check the delivery to make sure nothing is missing. Compare it against the shipping
documents and your order.
• Pack the product for storage and transportation in such a way that it is reliably protected
against impact and moisture. The original packaging offers the best protection.
Furthermore, the permitted ambient conditions must also be observed (see "Technical
data").
• If you have any questions, contact your supplier or your local sales center.
2.2
Product identification
2.2.1
Nameplate
The nameplate contains the following information:
• Manufacturer details
• Order code
• Extended order code
• Serial number
• Operating conditions
• Safety information symbols
Compare the order code on the nameplate with your order.
2.2.2
Identifying the product
The order code and serial number of your device can be found in the following locations:
• On the nameplate
• In the shipping documents
To discover what product version you have, enter the order code on the nameplate into
the search screen at the following address:
www.products.endress.com/order-ident
2.3
Scope of delivery
The scope of delivery comprises:
• 1 Turbimax CUS52D sensor in the version ordered
• 1 Set of Operating Instructions BA01275C/07/EN
If you have any questions, contact your supplier or your local sales center.
2.4
Certificates and approvals
Declaration of Conformity
The product meets the requirements of the harmonized European standards.
As such, it complies with the legal specifications of the EC directives.
The manufacturer confirms successful testing of the product by affixing to it the 4 mark.
6
Endress+Hauser
Turbimax CUS52D
Installation
3
Installation
3.1
Dimensions
a0022048
Fig. 1:
Dimensions in mm (inch)
Compressed air cleaning
Consumption: 50 l/min (13.2 gal/min)
Pressure: 1.5 to 2 bar (22 to 30 psi)
Connection: 6/8 mm or 6.35 mm (¼")
a0022011
Fig. 2:
Endress+Hauser
CUS52D with compressed air cleaning
7
Installation
Turbimax CUS52D
a0022911
Fig. 3:
8
Calkit CUS52 solid state reference, dimensions in mm (inch)
Endress+Hauser
Turbimax CUS52D
Installation
3.2
Installation instructions
3.2.1
Measuring system
A complete measuring system consists of:
• Turbimax CUS52D turbidity sensor
• Liquiline CM44x multichannel transmitter
• Direct installation in a pipe connection (2" clamp) or
• Assembly:
– CUA252 flow assembly or
– CUA262 flow assembly or
– Flexdip CYA112 assembly and Flexdip CYH112 holder system or
– Retractable assembly, e.g. Cleanfit CUA451
a0022017
Fig. 4:
1
2
3
Direct installation in a pipe connection (2" clamp)
Liquiline CM44x multichannel transmitter
Turbimax CUS52D turbidity sensor
Flow direction
a0022018
Fig. 5:
1
2
3
4
Endress+Hauser
Measuring system with flow assembly CUA252
Liquiline CM44x multichannel transmitter
Turbimax CUS52D turbidity sensor
CUA252 flow assembly
Flow direction
9
Installation
Turbimax CUS52D
a0022019
Fig. 6:
1
2
3
4
Measuring system with flow assembly CUA262
Liquiline CM44x multichannel transmitter
Turbimax CUS52D turbidity sensor
CUA262 flow assembly
Flow direction
a0022229
Fig. 7:
1
2
3
4
10
Measuring system with retractable assembly
Turbimax CUS52D turbidity sensor
Liquiline CM44x multichannel transmitter
Cleanfit CUA451 retractable assembly
Flow direction
Endress+Hauser
Turbimax CUS52D
Installation
a0022020
Fig. 8:
1
2
3
Measuring system with immersion assembly
Liquiline CM44x multichannel transmitter
Weather protection cover
Flexdip CYH112 holder system
4
5
Flexdip CYA112 assembly
Turbimax CUS52D turbidity sensor
a0022023
Fig. 9:
1
2
3
Endress+Hauser
Measuring system with immersion assembly on chain holder system
Flexdip CYH112 holder system
Liquiline CM44x multichannel transmitter
Weather protection cover
4
5
Flexdip CYA112 assembly
Turbimax CUS52D turbidity sensor
11
Installation
Turbimax CUS52D
3.3
Installation examples
3.3.1
Installation options
• with clamp connection
• with Flowfit CUA252 flow assembly
• with Flowfit CUA262 flow assembly
• with Cleanfit CUA451 retractable assembly
• with "E" or "S" flow assembly from CUS31
• with Flexdip CYA112 immersion assembly and Flexdip CYH112 holder system
The installation angle is 90°.
The arrow points in the direction of flow.
The optical windows in the sensor must be aligned
against the direction of flow.
A weld-in adapter is available as an accessory for
installation (see "Accessories" section).
a0022032
Fig. 10:
Installation with 2" clamp connection
>
10
0
(3
.9
4)
60 ... 64 (2.36 ... 2.52)
36
(1.42)
a0022844
Fig. 11:
Pipe connection with weld-in adapter, dimensions in mm (inch)
The installation angle is 90°.
The arrow points in the direction of flow.
The optical windows in the sensor must be aligned
against the direction of flow.
a0022034
Fig. 12:
12
Installation with flow assembly CUA252
Endress+Hauser
Turbimax CUS52D
Installation
The installation angle is 90°.
The arrow points in the direction of flow.
The optical windows in the sensor must be aligned
against the direction of flow.
a0022281
Fig. 13:
Installation with flow assembly CUA262
The installation angle is 90°.
The arrow points in the direction of flow.
The optical windows in the sensor must be aligned
against the direction of flow.
For manual insertion/retraction of the assembly the
medium pressure may not exceed 2 bar (29 psi).
a0022285
Fig. 14:
Installation with retractable assembly
CUA451
The installation angle is 90°.
The alignment of the sensor depends on the
medium. In the case of media with little outgassing
and with a tendency to form deposits, the optical
windows face downwards (as shown in the graphic).
In the case of media with strong outgassing and with
no tendency to form deposits, the optical windows
face upwards.
The installation marking on the sensor helps the
user align the sensor.
a0022687
Fig. 15:
Installation in CUS31-S flow assembly
1
a0022852
Fig. 16:
1
Endress+Hauser
Installation marking for sensor alignment in the CUS31-S flow assembly
Installation marking
13
Installation
Turbimax CUS52D
The arrow points in the direction of flow.
The installation angle is 0°.
If you are using the sensor in open basins, the sensor
must be installed in such a way that bubbles cannot
accumulate on it.
a0022033
Fig. 17:
14
Installation with immersion assembly
Endress+Hauser
Turbimax CUS52D
Installation
3.3.2
Pipe installation
3
3
5
4
1
2
2
a0022045
Fig. 18:
Orientation and positions
• Install the sensor in places with uniform flow conditions.
• The best installation location is in the ascending pipe (pos. 1). Installation is also possible
in the horizontal pipe (pos. 5).
• Do not install in places where air spaces or bubbles occur (pos. 3) or where sedimentation
may occur (pos. 2).
• Avoid installation in the down pipe (pos. 4).
• Avoid fittings downstream from pressure reduction stages which may cause outgassing.
Wall effects:
Backscattering on the pipe wall may result in the distortion of measurement values in the
case of turbidity values < 200 FNU. Assembly adjustment is recommended.
Black plastic pipes with diameter > DN 60 exhibit hardly any wall effects (<0.05 FNU). For
this reason, the use of black plastic pipes is recommended.
Endress+Hauser
15
Installation
Turbimax CUS52D
Additional information on avoiding wall effects:
a0022249
Fig. 19:
Orientations for pipes and assemblies
• Install the sensor in such a way that the light beam is not reflected (pos. 6).
• Avoid sudden changes in cross-section (pos. 9). Changes in cross-section should be
gradual and located as far away as possible from the sensor (pos. 10).
• Do not install the sensor directly downstream from a bend (pos. 7). Instead position it as
far away as possible from the bend (pos. 8).
• When using reflective materials (e.g. stainless steel), the pipe diameter must be at least
100 mm (4"). An assembly adjustment onsite is recommended.
• Pipes made of stainless steel with diameter >DN 300 exhibit hardly any wall effects.
16
Endress+Hauser
Turbimax CUS52D
Installation
3.3.3
Immersion operation
a0022020
Fig. 20:
1
2
3
Measuring system with immersion assembly
Flexdip CYH112 holder system
Liquiline CM44x multichannel transmitter
Weather protection cover
4
5
Flexdip CYA112 assembly
Turbimax CUS52D turbidity sensor
Weather protection cover
This type of installation is particularly suitable for strong or turbulent flow (>0.5 m/s
(1.6 ft/s)) in basins or channels.
Endress+Hauser
17
Installation
Turbimax CUS52D
Compressed air cleaning
a0022713
Fig. 21:
Installation marks 1 to 3
a0022715
Fig. 22:
Mounting position
Mount the compressed air cleaning system as follows:
1.
Fit the compressed air cleaning system on the sensor ( å 22).
2.
Position the securing ring of the compressed air cleaning system between installation
marks 2 and 3 ( å 21)
3.
Using a 4 mm Allen key, tighten the securing screw of the compressed air cleaning
system slightly so that the compressed air cleaning system can still be rotated.
4.
Turn the compressed air cleaning system so that the slit on the black ring is on
installation mark 1 ( å 21). This way the nozzle is offset by 20° when blowing air at
the optical windows.
5.
Tighten the securing screw.
6.
Fit the compressed air hose on the hose connection.
3.4
Post-installation check
‣ Sensor and cable undamaged?
‣ Is the orientation correct?
‣ Is the sensor installed in an assembly and not suspended freely from a cable?
18
Endress+Hauser
Turbimax CUS52D
Electrical connection
4
!
Electrical connection
WARNING
The device is live!
Incorrect connection may result in injury or death
‣ The electrical connection must only be established by an electrical technician.
‣ The electrical technician must have read and understood these Operating Instructions
and must follow the instructions they contain.
‣ Prior to commencing the connection work, make sure no voltage is applied at any of the
cables.
4.1
Connecting to the transmitter
The sensor is connected to the Liquiline CM442 transmitter as follows:
• With the M12 plug (version: fixed cable, M12 plug) or
• With the fixed cable connected to the terminal strips (version: fixed cable, end sleeves):
a0012460
Fig. 23:
Sensor connection
The maximum cable length is 100 m (328 ft).
4.2
Post-connection check
Device condition and specifications
Notes
Is damage visible on the outside of the sensor, assembly, junction box
or cable?
Visual inspection
Electrical connection
Notes
Does the transmitter supply voltage match the specifications on the
nameplate?
Are the installed cables strain-relieved and not twisted?
Is the cable type route completely isolated at the mounting location?
Power cables / signal cables
Are the power supply and signal cables connected correctly?
Use the transmitter wiring diagram
Are the cable cores stripped sufficiently and installed correctly in the
terminal?
Check if seated correctly (pull gently)
Are all the screw terminals firmly tightened?
Tighten
Are all the cable entries installed, firmly tightened and leak-tight?
For lateral cable entries, make sure the
cable loops downwards to ensure water
Are all the cable entries mounted on the side or pointing downwards?
can drain off.
Endress+Hauser
19
Device description
Turbimax CUS52D
5
Device description
5.1
Sensor design
The sensor is suitable for all drinking water and process water applications.
The sensor is designed as a 40 mm sensor that can be operated directly and completely in
the process without the need for further sampling (in situ).
All the necessary modules are contained in the sensor:
• Power supply
• Light sources
• Detectors detect the measuring signals, digitize them and process them to form a
measured value.
• The sensor microcontroller is responsible for controlling the internal processes and
transmitting the data.
All the data - including the calibration data - are stored in the sensor. The sensor can thus
be precalibrated and used at a measuring point, calibrated externally, or used for several
measuring points with different calibrations.
a0022012
Fig. 24:
1
2
5.2
Arrangement of light source and light receiver
Light receiver
Light source
Measuring principle
The sensor works using the 90° light scattering principle in accordance with ISO 7027 and
meets all the requirements of this standard (no divergence and a maximum divergence of
1.5°). The ISO 7027 standard is obligatory for turbidity measurements in the drinking water
sector.
90° ± 2.5°
20° - 30°
a0022230
Fig. 25:
Measurement in accordance with ISO 7027
Measurement is done using a wavelength of 860 nm.
20
Endress+Hauser
Turbimax CUS52D
Device description
5.3
Applications
The formazin factory calibration is used as the basis for precalibrating additional
applications and optimizing them for the different media characteristics.
Application: water
Recommended
operational range
Max. display range
Factory calibration for
formazin
0.000 to 4000 FNU
0.000 to 9999 FNU
Application: Kaolin
0 to 600 mg/l
0 to 3 g/l
Application: PSL
0 to 500
0 to 3000
Application: diatomite
0 to 2200 mg/l
0 to 10 g/l
To adapt to a specific application, customer calibrations can be carried out with up to 6
points.
Endress+Hauser
21
Device description
Turbimax CUS52D
5.4
Calibration
The sensor is precalibrated on leaving the factory. As such, it can be used in a wide range of
applications without the need for additional calibration.
In addition the sensor offers a variety of ways to adapt the measurement to the particular
application:
• Assembly adjustment (compensation of wall effects in pipes and assemblies)
• Calibration or adjustment (1 to 6 points)
• Enter a factor (multiply the measured values by a constant factor)
• Enter an offset (add/subtract a constant value to/from the measured values)
• Duplicate the factory calibration data records
a0022718-en
Fig. 26:
Flowchart
To use the "Offset", "Factor" or "Assembly adjustment" functions, a new data record must
first be created using 1 to 6-point calibration or by duplicating a factory data record.
22
Endress+Hauser
Turbimax CUS52D
Device description
5.4.1
Assembly adjustment
Both the optical design of the CUS52D sensor and the CUA252 and CUA262 flow assemblies
are optimized to minimize measured errors caused by wall effects in assemblies or pipes
(measured error in CUA252 < 0.02 FNU).
Using the "Assembly adjustment" function, it is possible to automatically compensate for the
remaining measured errors caused by wall effects. The programmed function is based on
formazin measurements and therefore may require subsequent calibration to adjust the
measurement to the corresponding application or medium.
The following types of assembly adjustment are available:
Adjustment
Description
PE 100
Adjustment to Flowfit CUA 252 flow assembly
1.4404 (AISI 316 L)
Adjustment to CUA 262 flow assembly
Standard customer adjustment
Adjustment to any pipes / assemblies
Specialist customer adjustment
Adjustment only recommended for E+H Service staff
• If the "PE100" and "1.4404/316L" assembly adjustment options are selected, no other
parameters must be specified.
• The "Standard customer adjustment" option requires the user to enter additional
parameters (material, surface and internal diameter).
5.4.2
Calibration / adjustment
The factory calibrations of the individual applications (Formazin, Kaolin, PSL, Diatomite) are
each based on 20 calibration points.
Apart from the uneditable factory calibrations, the sensor contains six additional data
records for saving process calibrations or for adaptation to the corresponding measuring
point (application).
5.4.3
Selecting the applications
On the CM44x, select the appropriate application for your field of application during initial
commissioning or calibration.
Model name
Application
Unit
Formazin
Drinking water, process water
FNU; NTU; TE/F; EBC; ASBG
Kaolin
Drinking water, filtratable substances, process
water
mg/l; g/l; ppm
PSL
The calibration standard commonly used in Japan
for drinking water turbidity
Diatomite
Mineral-based solids (sand)
(dough)
mg/l; g/l; ppm
1 to 6 points can be calibrated for all applications.
Endress+Hauser
23
Device description
Turbimax CUS52D
5.4.4
Single-point and multipoint calibration
• Before a calibration, rinse the system until all air pockets and fouling are removed.
• You can edit both the actual values and the target values in the calibration table (right and
left column)
• You can also add additional pairs of calibration values without measuring in a medium.
• Lines interpolate between the calibration points.
• When factory calibration data records are duplicated, the value pair
1000/1000 is automatically generated to map the factory data record 1:1 to the
duplicated record. If you perform a single-point or multipoint calibration after duplicating
the data record, you must delete this value pair (1000/1000) from the calibration table.
Single-point calibration
Application example:
The sensor measured value in an application deviates from the laboratory value. This is
corrected by a single-point calibration.
a0022722-en
Fig. 27:
1
2
Example for single-point calibration
Factory calibration
Application calibration
Proceed as follows:
1.
Select a data record.
2.
Set a calibration point in the medium or enter the target sample value (laboratory
value).
In our example ( å 27) these are the values 600 FNU (measured value) and 400 FNU
(target sample value). The gain for the application calibration changes accordingly.
24
Measured value
Target sample value
600
400
Endress+Hauser
Turbimax CUS52D
Device description
Two-point calibration
Application example:
Deviations in the measured value must be compensated for at two different points in an
application (e.g. maximum value and minimum value of the application). This aims to ensure
a maximum level of accuracy between these two extremes.
a0022723-en
Fig. 28:
1
2
Example for two-point calibration
Factory calibration
Application calibration
Proceed as follows:
1.
Select a data record.
2.
Set two different calibration points in the medium or enter the target sample values for
your application range.
In our example ( å 28) these are the values 300 FNU (measured value) and 400 FNU
(target sample value) as well as 800 FNU / 600 FNU. The gain and offset for the application
calibration change accordingly.
Measured value
Target sample value
300
400
800
600
The gain outside the selected operational range is determined by the adjacent
operational range (gray line).
The characteristic must increase monotonically.
Endress+Hauser
25
Device description
Turbimax CUS52D
Three-point calibration
a0022724-en
Fig. 29:
1
2
Example for three-point calibration
Factory calibration
Application calibration
Proceed as follows:
1.
Select a data record.
2.
Set three different calibration points in the medium or enter the measured values and
target sample values for your application range.
In our example ( å 29) these are 300 / 400, 600 / 500 and 800 / 800. The gain and
offset for the application calibration change accordingly.
Measured value
Target sample value
300
400
600
500
800
800
The gain outside the selected operational range is determined by the adjacent
operational range (gray line).
The characteristic must increase monotonically.
26
Endress+Hauser
Turbimax CUS52D
Device description
Four-point calibration
a0022725-en
Fig. 30:
1
2
Example for four-point calibration
Factory calibration
Application calibration
Proceed as follows:
1.
Select a data record.
2.
Set four different calibration points in the medium or enter the measured values and
target sample values for your application range.
In our example ( å 30) these are 200 / 200, 300 / 400, 600 / 500 and 800 / 800.
The gain and offset for the application calibration change accordingly.
Measured value
Target sample value
200
200
300
400
600
500
800
800
The gain outside the selected operational range is determined by the adjacent
operational range (gray line).
The characteristic must increase monotonically.
Endress+Hauser
27
Device description
Turbimax CUS52D
Calibration example for filter monitoring
Application example:
If a threshold is exceeded, the measured value is set to a maximum value regardless of the
actual turbidity.
a0022719-en
Fig. 31:
Example for filter monitoring
Values for example above ( å 31):
28
Measured value
Target sample value
0
0
100
100
101
1000
1000
1001
Endress+Hauser
Turbimax CUS52D
Device description
5.4.5
Factor
With the "Factor" function, the measured values are multiplied by a constant factor. The
functionality corresponds to that of a single-point calibration.
Example
This type of adjustment can be selected if the measured values were compared to the
laboratory values over an extended period and all the values are too low by a constant factor
of 10%, for instance.
The values are adjusted by entering the factor 1.1.
Application calibration
FNU
2000
1500
1000
500
500
1000
1500
2000
FNU
Factory calibration
a0022733-en
Fig. 32:
Example for factor
To use the "Factor" function, a new data record must first be created using 1 to 6-point
calibration or by duplicating a factory data record.
5.4.6
Offset
With the "Offset" function, the measured values are offset by a constant amount (added or
subtracted).
Application calibration
FNU
2000
1500
1000
500
500
1000
1500
2000
FNU
Factory calibration
a0022734-en
Fig. 33:
Example for offset
To use the "Offset" function, a new data record must first be created using 1 to 6-point
calibration or by duplicating a factory data record.
Endress+Hauser
29
Device description
Turbimax CUS52D
5.5
Cyclic cleaning
Compressed air is most suitable for cyclic cleaning in open basins and channels. The optional
cleaning unit is either ready-supplied or can be retrofitted, and is fitted on the sensor head.
It operates at a rate of 50 l/min (13.2 US gal/min)
The following settings are recommended for the cleaning unit:
Type of fouling
Cleaning interval
Cleaning duration
Severe fouling with rapid buildup
5 min
10 s
Low degree of fouling
10 min
10 s
The CYR52 ultrasonic cleaning system is suitable for cyclic cleaning in pipes and assemblies.
The (retrofittable) cleaning unit can be mounted on the CUA252, CUA262 flow assemblies,
the flow assembly of CUS31 or on any customer pipe.
The following cleaning settings are recommended to prevent the ultrasonic transducer from
overheating:
Cleaning duration
Cleaning interval
5.6
max. 5 seconds
min. 5 minutes
Signal filter
Optical turbidity measurements have a low signal-to-noise ratio particularly in the low
turbidity range. In addition, disturbance variables can occur as a result of air bubbles, fouling
etc. However a high level of damping affects the sensitivity of the measured value required
in applications.
For this reason, the sensor is fitted with an internal signal filter function in order to adapt
the measurement flexibly to different measuring requirements. The following filter settings
are possible:
30
Measured value
filter
Description
Weak
Low filtering, high sensitivity, fast response time (2 sec.) to changes
Normal (default)
Medium filtering, 10 sec. response time
Strong
Strong filtering, low sensitivity, slow response to changes (25 sec.)
Specialist
This menu is designed for the Endress+Hauser Service Team.
Endress+Hauser
Turbimax CUS52D
Device description
5.7
Calkit CUS52 solid state reference
The function and accuracy of the CUS52D sensor can be checked with the Calkit CUS52 solid
state references. During factory calibration, every solid state reference is aligned with a
special CUS52D sensor and can only be used with this sensor. Therefore the solid state
reference and the CUS52D sensor are married (permanently assigned) to one another.
The following solid state references are available:
• 5 FNU (NTU)
• 20 FNU (NTU)
• 50 FNU (NTU)
The reference value indicated on the solid state reference is reproduced with an accuracy of
±10% when the sensor is operating correctly.
Function check using Calkit CUS52
a0022914
Fig. 34:
Fitting the solid state reference on the sensor
Preparatory steps:
1.
Clean the sensor (see the "Cleaning the sensor" section).
2.
Fix the sensor in place (e.g. with a laboratory stand).
3.
With the solid state reference turned slightly ( å 34, B), fit it gently on the sensor
(C) and
4.
then allow the solid state reference to slide into the final position (D).
Function check:
1.
Enable the factory calibration on the transmitter.
2.
Read off the measured value at the transmitter (depending on the signal filter settings
it can take 2 to 25 seconds until the correct measured value appears).
3.
Compare the measured value with the reference value on the Calkit52. The sensor is
working perfectly if the value deviates by less than 10%.
If you activate a calibration data record other measured values will result. Therefore,
always select the factory calibration (formazin) when checking the function with Calkit
solid state references.
Endress+Hauser
31
Diagnostics and troubleshooting
Turbimax CUS52D
6
Diagnostics and troubleshooting
You must take the entire measuring point into account when troubleshooting:
• Transmitter
• Electrical connections and cables
• Assembly
• Sensor
The possible causes of error indicated in the table below primarily refer to the sensor.
Problem
Check
Remedial measures
Nothing displayed, no
reaction from the sensor
Power supplied to the transmitter?
Sensor connected correctly?
Buildup on optical windows?
Connect the mains voltage
Connect sensor correctly
Clean sensor
Display value too high or Buildup on optical windows?
too low
Sensor calibrated?
Clean
Calibrate
Display value fluctuates
a lot
Select other mounting location
Adjust the measured value filter
Check mounting location.
Please observe the troubleshooting instructions provided in the transmitter operating
manual. Examine the transmitter if necessary.
32
Endress+Hauser
Turbimax CUS52D
Maintenance
7
Maintenance
You have to perform maintenance tasks at regular intervals.
We recommend setting the maintenance times in advance in an operations journal or log.
The maintenance cycle primarily depends on the following:
• The facility
• The installation conditions
• The medium being measured
!
CAUTION
Risk of injury due to acid or medium
‣ Switch off the cleaning unit before removing the sensor from the medium.
7.1
Cleaning the sensor
Sensor fouling can affect the measurement results and even cause a malfunction.
The sensor must be cleaned at regular intervals to ensure reliable measurement results. The
frequency and intensity of the cleaning process depends on the medium.
Clean the sensor:
• As specified in the maintenance schedule
• Before every calibration
• Before returning the sensor for repair
Type of fouling
Cleaning measure
Lime deposits
Immerse the sensor in 1-5 % hydrochloric acid (for a
few minutes).
Dirt particles on the optical windows
Use a cleaning cloth to clean the optical windows.
You must rinse the sensor thoroughly with water after cleaning.
Endress+Hauser
33
Repair
Turbimax CUS52D
8
Repair
8.1
Spare parts kits
Order number
Description
71247510
Cable assembly, ferrules, 3 m cable length
71247510
Cable assembly, ferrules, 7 m cable length
71247510
Cable assembly, ferrules, 15 m cable length
71247510
Cable assembly, M12 plug, 3 m cable length
71247510
Cable assembly, M12 plug, 7 m cable length
71247510
Cable assembly, M12 plug, 15 m cable length
Detailed information on the spare parts kits is available in the "Spare Part Finding Tool",
which can be accessed on the Web at:
www.products.endress.com/spareparts_consumables
8.2
Return
The product must be returned if repairs or a factory calibration are required, or if the wrong
product has been ordered or delivered. According to legal regulations Endress+Hauser, as an
ISO-certified company, is required to follow certain procedures when handling returned
products that are in contact with the medium.
To ensure swift, safe and professional device returns:
Visit our website to obtain information about the return procedure and basic conditions:
www.services.endress.com/return-material
8.3
Disposal
The device contains electronic components and must therefore be disposed of in accordance
with regulations on the disposal of electronic waste.
Please observe local regulations.
34
Endress+Hauser
Turbimax CUS52D
Accessories
9
Accessories
9.1
Assemblies
Flowfit CUA252 flow assembly
• For CUS52D
• Order as per product structure (--> Online Configurator, www.products.endress.com/
cua252)
• Technical Information TI01139C/07/EN
Flowfit CUA262 flow assembly
• For CUS52D
• Order as per product structure (--> Online Configurator, www.products.endress.com/
cua262)
• Technical Information TI01152C/07/EN
Flexdip CYA112 wastewater assembly
• Modular assembly system for sensors in open basins, channels and tanks
• PVC and stainless steel version
• Order as per product structure (--> Online Configurator, www.products.endress.com/
cya112)
• Technical Information TI00432C/07/EN
Cleanfit CUA451 retractable assembly
• Manual retractable assembly made of stainless steel with ball valve shut-off for turbidity
sensors
• Order as per product structure (--> Online Configurator, www.products.endress.com/
cua451)
• Technical Information TI00369C/07/EN
9.2
Mounting material
Weld-in adapter for clamp connection DN 50
• Material: 1.4404 (AISI 316 L)
• Wall thickness 1.5 mm
• Order number: 71242201
36 (1.42)
30 (1.18)
Ø 53 (2.09)
Ø 64 (2.52)
a0022838
Fig. 35:
Endress+Hauser
Dimensions in mm (inch)
35
Accessories
Turbimax CUS52D
9.3
Holder
Flexdip CYH112 holder system for Flexdip CYA112 water and wastewater assemblies
• Modular holder system for sensors and assemblies in open basins, channels and tanks
• The holder system can be secured in any way, be it on the floor, the cap stone, the wall or
directly on a railing.
• Stainless steel version
• Order as per product structure (--> Online Configurator: www.products.endress.com/
cyh112)
• Technical Information TI00430C/07/EN
9.4
Compressed air cleaning
Compressed air cleaning for CUS52D
• Connection: 6 mm
• Materials: PE black
• Order no.: 71242026
a0022052
Fig. 36:
X
Compressed air cleaning, dimensions in mm (inch)
6 mm hose nozzle
Compressor
• For compressed air cleaning
• 230 V AC order no. 71072583
• 115 V AC order no. 71194623
9.5
Ultrasonic cleaning
Ultrasonic cleaning system CYR52
• For attachment to assemblies and pipes
• Order as per product structure (--> Online Configurator, www.products.endress.com/
cyr52)
• Technical Information TI01153C/07/EN
36
Endress+Hauser
Turbimax CUS52D
Accessories
9.6
Bubble trap
1
2
3
12 (3/8")
12 (3/8")
Bubble trap
• For sensor CUS52D
• Process pressure: up to 3 bar (43.5 psi)
• Process temperature: 0 to 50 °C (32 to 122 °F)
• Adapter to D 12 with connection for vent line (top connection on CUA252), is included in
the delivery.
• Orifice plates for the following volume flow rates:
– < 60 l/h (15.8 gal/hr)
– 60 to 100 l/h (15.8 to 26.4 gal/hr)
– > 100 l/h (26.4 gal/hr)
• The vent line is fitted with a PVC hose, hose check valve and a Luer lock adapter.
• Order number, suitable for assembly CUA252: 71242170
• Order number, suitable for assembly S of CUS31: 71247364
254 (10.0)
a0022631
Fig. 37:
1
2
3
9.7
Bubble trap, dimensions in mm (inch)
Inlet for medium (without hose system)
Outlet for bubbles (hose system is included in scope of supply)
Outlet for medium (without hose system)
Calibration set
CUY52 calibration set
• For CUS52D
• Order as per product structure (--> Online Configurator, www.products.endress.com/
cuy52)
• Technical Information TI01154C/07/EN
9.8
Transmitter
Liquiline CM44x/CM44xR
• Multichannel transmitter for connecting digital sensors with Memosens technology
• Field device (CM44x) or DIN rail device (CM44xR)
• Power supply 100 to 230 V AC, 24 V AC/DC
• Universally extensible
• Slot for SD card
• Order as per product structure (--> Configurator on product page)
• Technical Information TI00444C/07/EN (CM44x) or TI01112C/07/EN (CM44xR)
Endress+Hauser
37
Technical data
Turbimax CUS52D
10
Technical data
10.1
Input
Measured variables
Turbidity
Temperature
Measuring ranges
CUS52D
Turbidity
Temperature
0.000 to 4000 FNU
Display range up to 9999 FNU
-20 to +85 °C (-4 to +185 °F)
Application
Formazin
For measuring range up to 10 FNU, the sensor exhibits a detection limit (LOD) of 0.0015
FNU (measured in accordance with ISO 15839).
38
Endress+Hauser
Turbimax CUS52D
Technical data
10.2
Performance characteristics
Measured error
2% ±0.01 FNU; reference: factory calibration
Wavelength
860 nm
Conformity
Determining turbidity in accordance with ISO 7027
Factory calibration
The sensor has been calibrated in the factory for "formazin" applications.
Basis: internal 20-point characteristic curve
Applications
The formazin factory calibration is used as the basis for precalibrating additional
applications and optimizing them for the different media characteristics.
Application: water
Recommended operating
ranges
Max. display range
Factory calibration for
formazin
0.000 to 4000 FNU
0.000 to 9999 FNU
Application: Kaolin
0 to 600 mg/l
0 to 3 g/l
Application: PSL
0 to 500
0 to 3000
Application: diatomite
0 to 2200 mg/l
0 to 10 g/l
To adapt to a specific application, customer calibrations can be carried out with up to 6
points.
Drift
Detection limits
Working on the basis of electronic controls, the sensor is largely free of drifts.
Application
Measuring range
Detection limit
Formazin
0 to 10 FNU (ISO 15839)
0.0015 FNU
Response time
>1 sec, adjustable
Repeatability
< 0.5 % of measured value (measuring range : 0 to 10 FNU)
Endress+Hauser
39
Technical data
Turbimax CUS52D
10.3
Environment
Ambient temperature
range
-20 to 60 °C (-4 to 140 °F)
Storage temperature
-20 to 70 °C (-4 to 158 °F)
Degree of protection
IP 68 (test condition: 1.8 m (5.91 ft) water column over 20 days, 1 mol/l KCl)
10.4
Process
Process temperature
-20 to 85 °C (-4 to 185 °F)
Process pressure
0.5 to 10 bar (7 to 145 psi) absolute
Minimum flow
No minimum flow necessary.
For solids which have a tendency to form deposits, ensure that sufficient mixing is
performed.
10.5
Mechanical construction
Dimensions
See "Installation conditions"
Weight
Sensor with 7 m cable
With clamp
Without clamp
approx. 1.56 kg
approx. 1.48 kg
Materials
Sensor
Optical windows
O-rings
Process connections
G1 and NPT ¾''
2" clamp (depends on sensor version)/ DIN 32676
40
Stainless steel 1.4404 (AISI 316 L)
Sapphire
EPDM
Endress+Hauser
Turbimax CUS52D
Endress+Hauser
Technical data
41
Turbimax CUS52D
Index
A
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Ambient temperature range . . . . . . . . . . . . . . . . . . . . . . 40
Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21, 39
Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Assembly adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
B
Bubble trap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
C
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Calibration set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Connection
Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cyclic cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
22
37
33
19
30
D
Damping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Designated use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
E
Electrical technician . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Electromagnetic compatibility . . . . . . . . . . . . . . . . . . . . . . . 4
F
Factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Factory calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Filter monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
H
Holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
I
Immersion operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Input
Measured variables . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Measuring ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Installation examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Installation instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
M
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Measured error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Measured variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Measuring ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Minimum flow. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Mounting material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Multipoint calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
N
Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
O
Occupational safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Operational safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
P
Performance characteristics
Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Measured error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Wavelength . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Pipe installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Post-connection check . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Post-installation check . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Process pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Process temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Product identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Product safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
R
Requirements for personnel . . . . . . . . . . . . . . . . . . . . . . . . 4
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
S
Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Selecting the applications . . . . . . . . . . . . . . . . . . . . . . . . . 23
Sensor design. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Signal filter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Single-point calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Solid state reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Spare parts kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Storage temperature. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
T
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Transmitter connection . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
U
Ultrasonic cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
42
Endress+Hauser
Turbimax CUS52D
W
Wavelength . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19