Teledyne Analytical Instruments OPERATING INSTRUCTIONS FOR Model 3000TA Trace Oxygen Analyzer
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Trace Oxygen Analyzer OPERATING INSTRUCTIONS FOR Model 3000TA Trace Oxygen Analyzer DANGER HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM. PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM. HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST FOR A TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED. ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING. BEFORE CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/ MANAGER. Teledyne Analytical Instruments P/N M62928 11/24/04 ECO: #03-0126 i Model 3000TA Copyright © 1999 Teledyne Analytical Instruments All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer language in whole or in part, in any form or by any means, whether it be electronic, mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of Teledyne Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 917491580. Warranty This equipment is sold subject to the mutual agreement that it is warranted by us free from defects of material and of construction, and that our liability shall be limited to replacing or repairing at our factory (without charge, except for transportation), or at customer plant at our option, any material or construction in which defects become apparent within one year from the date of shipment, except in cases where quotations or acknowledgements provide for a shorter period. Components manufactured by others bear the warranty of their manufacturer. This warranty does not cover defects caused by wear, accident, misuse, neglect or repairs other than those performed by Teledyne or an authorized service center. We assume no liability for direct or indirect damages of any kind and the purchaser by the acceptance of the equipment will assume all liability for any damage which may result from its use or misuse. We reserve the right to employ any suitable material in the manufacture of our apparatus, and to make any alterations in the dimensions, shape or weight of any parts, in so far as such alterations do not adversely affect our warranty. Important Notice This instrument provides measurement readings to its user, and serves as a tool by which valuable data can be gathered. The information provided by the instrument may assist the user in eliminating potential hazards caused by his process; however, it is essential that all personnel involved in the use of the instrument or its interface, with the process being measured, be properly trained in the process itself, as well as all instrumentation related to it. The safety of personnel is ultimately the responsibility of those who control process conditions. While this instrument may be able to provide early warning of imminent danger, it has no control over process conditions, and it can be misused. In particular, any alarm or control systems installed must be tested and understood, both as to how they operate and as to how they can be defeated. Any safeguards required such as locks, labels, or redundancy, must be provided by the user or specifically requested of Teledyne at the time the order is placed. Therefore, the purchaser must be aware of the hazardous process conditions. The purchaser is responsible for the training of personnel, for providing hazard warning methods and instrumentation per the appropriate standards, and for ensuring that hazard warning devices and instrumentation are maintained and operated properly. Teledyne Analytical Instruments, the manufacturer of this instrument, cannot accept responsibility for conditions beyond its knowledge and control. No statement expressed or implied by this document or any information disseminated by the manufacturer or its agents, is to be construed as a warranty of adequate safety control under the user’s process conditions. ii Teledyne Analytical Instruments Trace Oxygen Analyzer Specific Model Information The instrument for which this manual was supplied may incorporate one or more options not supplied in the standard instrument. Commonly available options are listed below, with check boxes. Any that are incorporated in the instrument for which this manual is supplied are indicated by a check mark in the box. Instrument Serial Number: _______________________ Options Included in the Instrument with the Above Serial Number: q 3000TA-C: In addition to all standard features, this model also has separate ports for zero and span gases, and builtin control valves. The internal valves are entirely under the control of the 3000TA electronics, to automatically switch between gases in synchronization with the analyzer’s operations q 19" Rack Mnt: The 19" Relay Rack Mount units are available with either one or two 3000 series analyzers installed in a standard 19" panel and ready to mount in a standard rack. Teledyne Analytical Instruments iii Model 3000TA Table of Contents 1 Introduction 1.1 1.2 1.3 1.4 1.5 1.6 1.7 Overview ........................................................................ 1-1 Typical Applications ....................................................... 1-1 Main Features of the Analyzer ....................................... 1-1 Model Designations ....................................................... 1-2 Front Panel (Operator Interface) .................................... 1-3 Recognizing Difference Between LCD & VFD ............... 1-5 Rear Panel (Equipment Interface) .................................. 1-5 2 Operational Theory 2.1 Introduction .................................................................... 2-1 2.2 Micro-Fuel Cell Sensor .................................................. 2-1 2.2.1 Principles of Operation ............................................ 2-1 2.2.2 Anatomy of a Micro-Fuel Cell .................................. 2-2 2.2.3 Electrochemical Reactions ...................................... 2-3 2.2.4 The Effect of Pressure ............................................. 2-4 2.2.5 Calibration Characteristics ...................................... 2-4 2.3 Sample System .............................................................. 2-5 2.4 Electronics and Signal Processing ................................ 2-7 3 Installation 3.1 Unpacking the Analyzer ................................................. 3-1 3.2 Mounting the Analyzer ................................................... 3-1 3.3 Rear Panel Connections ................................................ 3-2 3.3.1 Gas Connections ................................................... 3-3 3.3.2 Electrical Connections ........................................... 3-4 3.3.3 Remote Probe Connector ...................................... 3-9 3.4 Installing the Micro-Fuel Cell ........................................... 310 3.5 Testing the System ........................................................ 3-10 4 Operation 4.1 Introduction .................................................................... 4-1 4.2 Using the Data Entry and Function Buttons ................... 4-2 4.3 The System Function ..................................................... 4-3 4.3.1 Tracking O2 Readings during Calibration ............... 4-4 4.3.2 Setting up an Auto-Cal ........................................... 4-5 4.3.3 Password Protection .............................................. 4-6 4.3.3.1 Entering the Password ................................... 4-6 4.3.3.2 Installing or Changing the Password ............. 4-7 iv Teledyne Analytical Instruments Trace Oxygen Analyzer 4.3.4 Logout .................................................................... 4-9 4.3.5 System Self-Diagnostic Test .................................. 4-9 4.3.6 Version Screen ...................................................... 4-10 4.3.7 Showing Negative Oxygen Readings .................... 4-11 4.4 The Zero and Span Functions ....................................... 4-11 4.4.1 Zero Cal ................................................................. 4-12 4.4.1.1 Auto Mode Zeroing ................................................ 4-12 4.4.1.2 Manual Mode Zeroing .................................... 4-13 4.4.1.3 Cell Failure ..................................................... 4-13 4.4.2 Span Cal ................................................................ 4-14 4.4.2.1 Auto Mode Spanning ..................................... 4-14 4.4.2.2 Manual Mode Spanning ................................. 4-15 4.4.3 Span Failure........................................................... 4-16 4.5 The Alarms Function ...................................................... 4-16 4.6 The Range Function ...................................................... 4-18 4.6.1 Setting the Analog Output Ranges ........................ 4-19 4.6.2 Fixed Range Analysis ............................................ 4-20 4.7 The Analyze Function .................................................... 4-20 4.8 Signal Output ................................................................. 4-20 Maintenance 5.1 Routine Maintenance ..................................................... 5-1 5.2 Cell Replacement .......................................................... 5-1 5.2.1 Storing and Handling Replacement Cells ............... 5-1 5.2.2 When to Replace a Cell ........................................... 5-2 5.2.3 Removing the Micro-Fuel Cell ................................. 5-2 5.2.4 Installing a New Micro-Fuel Cell .............................. 5-4 5.2.5 Cell Warranty ........................................................... 5-4 5.3 Fuse Replacement ......................................................... 5-5 5.4 System Self Diagnostic Test........................................... 5-6 5.5 Major Internal Components ........................................... 5-6 5.6 Cleaning ......................................................................... 5-7 5.7 Troubleshooting.............................................................. 5-8 Appendix A-1 A-2 A-3 A-4 A-5 A-6 Model 3000TA Specifications ......................................... A-1 Recommended 2-Year Spare Parts List ......................... A-2 Drawing List ................................................................... A-3 19-Inch Relay Rack Panel Mount ................................... A-4 Application Notes on Restrictors, Pressures and Flow .. A-5 Material Safety Data Sheet ............................................ A-6 Teledyne Analytical Instruments v Model 3000TA DANGER COMBUSTIBLE GAS USAGE WARNING This is a general purpose instrument designed for usage in a nonhazardous area. It is the customer's responsibility to ensure safety especially when combustible gases are being analyzed since the potential of gas leaks always exist. The customer should ensure that the principles of operating of this equipment is well understood by the user. Misuse of this product in any manner, tampering with its components, or unauthorized substitution of any component may adversely affect the safety of this instrument. Since the use of this instrument is beyond the control of Teledyne, no responsibility by Teledyne, its affiliates, and agents for damage or injury from misuse or neglect of this equipment is implied or assumed. vi Teledyne Analytical Instruments Trace Oxygen Analyzer Introduction 1 Introduction 1.1 Overview The Teledyne Analytical Instruments Model 3000TA Trace Oxygen Analyzer is a versatile microprocessor-based instrument for detecting oxygen at the parts-per-million (ppm) level in a variety of gases. This manual covers the Model 3000TA General Purpose flush-panel and/or rack-mount units only. These units are for indoor use in a nonhazardous environment. 1.2 Typical Applications A few typical applications of the Model 3000TA are: • Monitoring inert gas blanketing • Air separation and liquefaction • Chemical reaction monitoring • Semiconductor manufacturing • Petrochemical process control • Quality assurance • Gas analysis certification. 1.3 Main Features of the Analyzer The Model 3000TA Trace Oxygen Analyzer is sophisticated yet simple to use. The main features of the analyzer include: • A 2-line alphanumeric display screen, driven by microprocessor electronics, that continuously prompts and informs the operator. • High resolution, accurate readings of oxygen content from low ppm levels through 25%. Large, bright, meter readout. • Stainless steel cell block. Teledyne Analytical Instruments 1-1 1 Introduction Model 3000TA • Advanced Micro-Fuel Cell, redesigned for trace analysis, has a one year warranty and an expected lifetime of two years. • Versatile analysis over a wide range of applications. • Microprocessor based electronics: 8-bit CMOS microprocessor with 32 kB RAM and 128 kB ROM. • Three user definable output ranges (from 0-10 ppm through 0250,000 ppm) allow best match to users process and equipment. • Air-calibration range for convenient spanning at 20.9 %. • Auto Ranging allows analyzer to automatically select the proper preset range for a given measurement. Manual override allows the user to lock onto a specific range of interest. • Two adjustable concentration alarms and a system failure alarm. • Extensive self-diagnostic testing, at startup and on demand, with continuous power-supply monitoring. • Two way RFI protection. • RS-232 serial digital port for use with a computer or other digital communication device. • Four analog outputs: two for measurement (0–1 V dc and Isolated 4–20 mA dc) and two for range identification. • Convenient and versatile, steel, flush-panel or rack-mountable case with slide-out electronics drawer. 1.4 Model Designations 3000TA: Standard model. 3000TA-C: In addition to all standard features, this model also has separate ports for zero and span gases, and built-in control valves. The internal valves are entirely under the control of the 3000TA electronics, to automatically switch between gases in synchronization with the analyzer’s operations. 1-2 Teledyne Analytical Instruments Trace Oxygen Analyzer 1.5 Introduction 1 Front Panel (Operator Interface) The standard 3000TA is housed in a rugged metal case with all controls and displays accessible from the front panel. See Figure 1-1. The front panel has thirteen buttons for operating the analyzer, a digital meter, an alphanumeric display, and a window for viewing the sample flowmeter. Function Keys: Six touch-sensitive membrane switches are used to change the specific function performed by the analyzer: Figure 1-1: Model 3000TA Front Panel • Analyze Perform analysis for oxygen content of a sample gas. • System Perform system-related tasks (described in detail in chapter 4, Operation.). • Span Span calibrate the analyzer. • Zero Zero calibrate the analyzer. • Alarms Set the alarm setpoints and attributes. • Range Set up the 3 user definable ranges for the instrument. Teledyne Analytical Instruments 1-3 1 Introduction Model 3000TA Data Entry Keys: Six touch-sensitive membrane switches are used to input data to the instrument via the alphanumeric VFD display: • Left & Right Arrows Select between functions currently displayed on the VFD screen. • Up & Down Arrows Increment or decrement values of functions currently displayed. • Enter • Escape Moves VFD display back to the previous screen in a series. If none remains, returns to the Analyze screen. Moves VFD display on to the next screen in a series. If none remains, returns to the Analyze screen. Digital Meter Display: The meter display is a LED device that produces large, bright, 7-segment numbers that are legible in any lighting. It produces a continuous readout from 0-10,000 ppm and then switches to a continuous percent readout from 1-25%. It is accurate across all analysis ranges without the discontinuity inherent in analog range switching. Alphanumeric Interface Screen: The VFD screen is an easy-to-use interface from operator to analyzer. It displays values, options, and messages that give the operator immediate feedback. Flowmeter: Monitors the flow of gas past the sensor. Readout is 0.2 to 2.4 standard liters per minute (SLPM). Standby Button: The Standby turns off the display and outputs, but circuitry is still operating. CAUTION: The power cable must be unplugged to fully disconnect power from the instrument. When chassis is exposed or when access door is open and power cable is connected, use extra care to avoid contact with live electrical circuits . Access Door: For access to the Micro-Fuel Cell, the front panel swings open when the latch in the upper right corner of the panel is pressed all the way in with a narrow gauge tool. Accessing the main circuit board requires unfastening rear panel screws and sliding the unit out of the case. 1.6 Recognizing Difference Between LCD & VFD LCD has GREEN background with BLACK characters. VFD has DARK background with GREEN characters. In the case of VFD - NO CONTRAST ADJUSTMENT IS NEEDED. 1-4 Teledyne Analytical Instruments Trace Oxygen Analyzer 1.7 Introduction 1 Rear Panel (Equipment Interface) The rear panel, shown in Figure 1-2, contains the gas and electrical connectors for external inlets and outlets. Some of those depicted are optional and may not appear on your instrument. The connectors are described briefly here and in detail in chapter 3 Installation. Figure 1-2: Model 3000TA Rear Panel • Power Connection Universal AC power source. • Gas Inlet and Outlet One inlet (must be externally valved) and one exhaust out. • Analog Outputs 0–1 V dc concentration plus 0-1 V dc range ID, and isolated 4–20 mA dc plus 4-20 mA dc range ID. • Alarm Connections 2 concentration alarms and 1 system alarm. • RS-232 Port Serial digital concentration signal output and control input. • Remote Probe Used in the 3000TA for controlling external solenoid valves only. • Remote Span/Zero Digital inputs allow external control of analyzer calibration. Teledyne Analytical Instruments 1-5 1 Introduction Model 3000TA • Calibration Contact To notify external equipment that instrument is being calibrated and readings are not monitoring sample. • Range ID Contacts Four separate, dedicated, range relay contacts. Low, Medium, High, Cal. • Network I/O Serial digital communications for local network access. For future expansion. Not implemented at this printing. Optional: • Calibration Gas Ports Separate fittings for zero, span and sample gas input, and internal valves for automatically switching the gases. Note: If you require highly accurate Auto-Cal timing, use external Auto-Cal control where possible. The internal clock in the Model 3000TA is accurate to 2-3 %. Accordingly, internally scheduled calibrations can vary 2-3 % per day. 1-6 Teledyne Analytical Instruments Trace Oxygen Analyzer Operational Theory 2 Operational Theory 2.1 Introduction The analyzer is composed of three subsystems: 1. Micro-Fuel Cell Sensor 2. Sample System 3. Electronic Signal Processing, Display and Control The sample system is designed to accept the sample gas and transport it through the analyzer without contaminating or altering the sample prior to analysis. The Micro-Fuel Cell is an electrochemical galvanic device that translates the amount of oxygen present in the sample into an electrical current. The electronic signal processing, display and control subsystem simplifies operation of the analyzer and accurately processes the sampled data. The microprocessor controls all signal processing, input/output and display functions for the analyzer. 2.2 Micro-Fuel Cell Sensor 2.2.1 Principles of Operation The oxygen sensor used in the Model 3000T series is a Micro-Fuel Cell designed and manufactured by Analytical Instruments. It is a sealed plastic disposable electrochemical transducer. The active components of the Micro-Fuel Cell are a cathode, an anode, and the 15% aqueous KOH electrolyte in which they are immersed. The cell converts the energy from a chemical reaction into an electrical current in an external electrical circuit. Its action is similar to that of a battery. There is, however, an important difference in the operation of a battery as compared to the Micro-Fuel Cell: In the battery, all reactants are stored within the cell, whereas in the Micro-Fuel Cell, one of the reactants Teledyne Analytical Instruments 2-1 2 Operational Theory Model 3000TA (oxygen) comes from outside the device as a constituent of the sample gas being analyzed. The Micro-Fuel Cell is therefore a hybrid between a battery and a true fuel cell. (All of the reactants are stored externally in a true fuel cell.) 2.2.2 Anatomy of a Micro-Fuel Cell The Micro-Fuel Cell is a cylinder only 1¼ inches in diameter and 1¼ inches thick. It is made of an extremely inert plastic, which can be placed confidently in practically any environment or sample stream. It is effectively sealed, although one end is permeable to oxygen in the sample gas. The other end of the cell is a contact plate consisting of two concentric foil rings. The rings mate with spring-loaded contacts in the sensor block assembly and provide the electrical connection to the rest of the analyzer. Figure 2-1 illustrates the external features. Figure 2-1: Micro-Fuel Cell Refer to Figure 2-2, Cross Section of a Micro-Fuel Cell, which illustrates the following internal description. Figure 2-2. Cross Section of a Micro-Fuel Cell (not to scale) 2-2 Teledyne Analytical Instruments Trace Oxygen Analyzer Operational Theory 2 At the top end of the cell is a diffusion membrane of Teflon, whose thickness is very accurately controlled. Beneath the diffusion membrane lies the oxygen sensing element—the cathode—with a surface area almost 4 cm2. The cathode has many perforations to ensure sufficient wetting of the upper surface with electrolyte, and it is plated with an inert metal. The anode structure is below the cathode. It is made of lead and has a proprietary design which is meant to maximize the amount of metal available for chemical reaction. At the rear of the cell, just below the anode structure, is a flexible membrane designed to accommodate the internal volume changes that occur throughout the life of the cell. This flexibility assures that the sensing membrane remains in its proper position, keeping the electrical output constant. The entire space between the diffusion membrane, above the cathode, and the flexible rear membrane, beneath the anode, is filled with electrolyte. Cathode and anode are submerged in this common pool. They each have a conductor connecting them to one of the external contact rings on the contact plate, which is on the bottom of the cell. 2.2.3 Electrochemical Reactions The sample gas diffuses through the Teflon membrane. Any oxygen in the sample gas is reduced on the surface of the cathode by the following HALF REACTION: O2 + 2H2O + 4e– → 4OH– (cathode) (Four electrons combine with one oxygen molecule—in the presence of water from the electrolyte—to produce four hydroxyl ions.) When the oxygen is reduced at the cathode, lead is simultaneously oxidized at the anode by the following HALF REACTION: Pb + 2OH– → Pb+2 + H2O + 2e– (anode) (Two electrons are transferred for each atom of lead that is oxidized. Therefore it takes two of the above anode reactions to balance one cathode reaction and transfer four electrons.) The electrons released at the surface of the anode flow to the cathode surface when an external electrical path is provided. The current is proportional to the amount of oxygen reaching the cathode. It is measured and used to determine the oxygen concentration in the gas mixture. Teledyne Analytical Instruments 2-3 2 Operational Theory Model 3000TA The overall reaction for the fuel cell is the SUM of the half reactions above, or: 2Pb + O2 → 2PbO (These reactions will hold as long as no gaseous components capable of oxidizing lead—such as iodine, bromine, chlorine and fluorine—are present in the sample.) The output of the fuel cell is limited by (1) the amount of oxygen in the cell at the time and (2) the amount of stored anode material. In the absence of oxygen, no current is generated. 2.2.4 The Effect of Pressure In order to state the amount of oxygen present in the sample in partsper-million or a percentage of the gas mixture, it is necessary that the sample diffuse into the cell under constant pressure. If the total pressure increases, the rate that oxygen reaches the cathode through the diffusing membrane will also increase. The electron transfer, and therefore the external current, will increase, even though the oxygen concentration of the sample has not changed. It is therefore important that the sample pressure at the fuel cell (usually vent pressure) remain relatively constant between calibrations. 2.2.5 Calibration Characteristics Given that the total pressure of the sample gas on the surface of the Micro-Fuel Cell input is constant, a convenient characteristic of the cell is that the current produced in an external circuit is directly proportional to the rate at which oxygen molecules reach the cathode, and this rate is directly proportional to the concentration of oxygen in the gaseous mixture. In other words it has a linear characteristic curve, as shown in Figure 2-3. Measuring circuits do not have to compensate for nonlinearities. In addition, since there is zero output in the absence oxygen, the characteristic curve has close to an absolute zero (within ± 1 ppm oxygen). In practical application, zeroing may still used to compensate for the combined zero offsets of the cell and the electronics. (The electronics is zeroed automatically when the instrument power is turned on.) 2-4 Teledyne Analytical Instruments Trace Oxygen Analyzer Operational Theory 2 Figure 2-3. Characteristic Input/Output Curve for a Micro-Fuel Cell 2.3 Sample System The sample system delivers gases to the Micro-Fuel Cell sensor from the analyzer rear panel inlet. Depending on the mode of operation either sample or calibration gas is delivered. The Model 3000TA sample system is designed and fabricated to ensure that the oxygen concentration of the gas is not altered as it travels through the sample system. The sample encounters almost no dead space. This minimizes residual gas pockets that can interfere with trace analysis. The sample system for the standard instrument incorporates ¼ inch tube fittings for sample inlet and outlet connections at the rear panel. For metric system installations, 6 mm adapters are supplied with each instrument to be used if needed. The sample or calibration gas flows through the system is monitored by a flowmeter downstream from the cell. Figure 2-4 shows the piping layout and flow diagram for the standard model. Teledyne Analytical Instruments 2-5 2 Operational Theory Model 3000TA Figure 2-4: Piping Layout and Flow Diagram for Standard Model Figure 2-5 is the flow diagram for the sampling system. In the standard instrument, calibration gases (zero and span) can be connected directly to the Sample In port by teeing to the port with appropriate valves. The shaded portion of the diagram shows the components added when the –C option is ordered. The valving is installed inside the 3000TA-C enclosure and is regulated by the instruments internal electronics. 2-6 Teledyne Analytical Instruments Trace Oxygen Analyzer Operational Theory 2 Figure 2-5: Flow Diagram 2.4 Electronics and Signal Processing The Model 3000TA Trace Oxygen Analyzer uses an 8031 microcontroller with 32 kB of RAM and 128 kB of ROM to control all signal processing, input/output, and display functions for the analyzer. System power is supplied from a universal power supply module designed to be compatible with any international power source. Figure 2-6 shows the location of the power supply and the main electronic PC boards. Figure 2-6: Electronic Component Location Inside the Model 3000TA Teledyne Analytical Instruments 2-7 2 Operational Theory Model 3000TA The signal processing electronics including the microprocessor, analog to digital, and digital to analog converters are located on the motherboard at the bottom of the case. The preamplifier board is mounted on top of the motherboard as shown in the figure. These boards are accessible after removing the back panel. Figure 2-7 is a block diagram of the Analyzer electronics. Figure 2-7: Block Diagram of the Model 3000TA Electronics 2-8 Teledyne Analytical Instruments Trace Oxygen Analyzer Operational Theory 2 In the presence of oxygen the cell generates a current. A current to voltage amplifier converts this current to a voltage, which is amplified in the second stage amplifier. The second stage amplifier also supplies temperature compensation for the oxygen sensor output. This amplifier circuit incorporates a thermistor, which is physically located in the cell block. The thermistor is a temperature dependent resistance that changes the gain of the amplifier in proportion to the temperature changes in the block. This change is inversely proportional to the change in the cell output due to the same temperature changes. The result is a signal that is temperature independent. The output from the second stage amplifier is sent to an 18 bit analog to digital converter controlled by the microprocessor. The digital concentration signal along with input from the control panel is processed by the microprocessor, and appropriate control signals are directed to the display, alarms and communications port. The same digital information is also sent to a 12 bit digital to analog converter that produces the 4-20 mA dc and the 0-1 V dc analog concentration signal outputs, and the analog range ID outputs. Signals from the power supply are also monitored, and through the microprocessor, the system failure alarm is activated if a malfunction is detected. Teledyne Analytical Instruments 2-9 2 Operational Theory 2-10 Teledyne Analytical Instruments Model 3000TA Trace Oxygen Analyzer Installation 3 Installation Installation of the Model 3000TA Analyzer includes: 1. 2. 3. 4. 5. 6. Unpacking Mounting Gas connections Electrical connections Installing the Micro-Fuel Cell Testing the system. 3.1 Unpacking the Analyzer The analyzer is shipped with all the materials you need to install and prepare the system for operation. Carefully unpack the analyzer and inspect it for damage. Immediately report any damage to the shipping agent. 3.2 Mounting the Analyzer The Model 3000TA is for indoor use in a general purpose area. It is NOT for hazardous environments of any type. The standard model is designed for flush panel mounting. Figure 3-1 is an illustration of the 3000TA standard front panel and mounting bezel. There are four mounting holes—one in each corner of the rigid frame. Drawing number D-62928, at the back of this manual, contains a panel cutout diagram. On special order, a 19" rack-mounting panel can be provided. For rack mounting, one or two 3000 series analyzers are flush-panel mounted on the rack panel. See Appendix for dimensions of the mounting panel. Teledyne Analytical Instruments 3-1 3 Installation Model 3000TA 6.7" 10" Figure 3-1: Front Panel of the Model 3000TA All operator controls are mounted on the control panel, which is hinged on the left edge and doubles as the door that provides access to the sensor and cell block inside the instrument. The door is spring loaded and will swing open when the button in the center of the latch (upper right corner) is pressed all the way in with a narrow gauge tool (less than 0.18 inch wide), such as a small hex wrench or screwdriver Allow clearance for the door to open in a 90-degree arc of radius 7.125 inches. See Figure 3-2. Figure 3-2: Required Front Door Clearance 3.3 Rear Panel Connections Figure 3-3 shows the Model 3000TA rear panel. There are ports for gas inlet and outlet, power, communication, and both digital and analog concentration output. The Zero In and Span In ports are not included on the standard model, but are available as options. 3-2 Teledyne Analytical Instruments Trace Oxygen Analyzer Installation 3 Figure 3-3: Rear Panel of the Model 3000TA 3.3.1 Gas Connections Before using this instrument, it should be determined if the unit will be used for pressurized service or vacuum service and low pressure applications. Inspect the restrictor kit that came with the unit. The kit consist of two restrictors and a union for 1/4” diameter tubing. Notice that the two 1 3/4” long, 1/4” diameter tubing are restrictors. It has an open end and a closed end with a small circular orifice. The restrictor without the blue sticker is for ;ow pressure and vacuum service. For high pressure (5 to 50 psig) applications, use the restrictor that has a blue sticker on the body. For pressurized service, use the restrictor without the blue dot and union from the restrictor kit and attach it to the Sample In port. The small circular orifice should face away from the back of the unit (against the direction of gas flow). Use the restrictor without the blue dot sticker in the same manner for low pressure applications (less than 5 psig). For vacuum service (5-10 in Hg), use the restrictor without the blue dot sticker and union but attach it to the Exhaust Out port. The small circular orifice should face toward the back of the unit (against the direction of gas flow). Remove the blue sticker from the restrictor before using. WARNING: Operating the unit without restrictors can cause damage to the micro-fuel cell. Teledyne Analytical Instruments 3-3 3 Installation Model 3000TA The unit is manufactured with 1/4 inch tube fittings, and 6 mm adapters are supplied for metric system installations. For a safe connection: 1. Insert the tube into the tube fitting, and finger-tighten the nut until the tubing cannot be rotated freely, by hand, in the fitting. (This may require an additional 1/8 turn beyond finger-tight.) 2. Hold the fitting body steady with a backup wrench, and with another wrench rotate the nut another 1-1/4 turns. SAMPLE IN: In the standard model, gas connections are made at the SAMPLE IN and EXHAUST OUT connections. Calibration gases must be Tee'd into the Sample inlet with appropriate valves. The gas pressure in should be reasonably regulated. Pressures between 2 and 50 psig are acceptable as long as the pressure, once established, will keep the front panel flowmeter reading in an acceptable range (0.1 to 2.4 SLPM). For non-pressurized sample or very low pressure, (2 psig or less) vacuum service pluming is recommended. Exact figures will depend on your process. If greater flow is required for improved response time, install a bypass in the sampling system upstream of the analyzer input. EXHAUST OUT: Exhaust connections must be consistent with the hazard level of the constituent gases. Check Local, State, and Federal laws, and ensure that the exhaust stream vents to an appropriately controlled area if required. ZERO IN and SPAN IN (Optional): These are additional ports for inputting span gas and zero gas. There are electrically operated valves inside for automatic switching between sample and calibration gases. These valves are completely under control of the 3000T Electronics. They can be externally controlled only indirectly through the Remote Cal Inputs, described below. Pressure, flow, and safety considerations are the same as prescribed for the SAMPLE IN inlet, above. 3.3.2 Electrical Connections For safe connections, ensure that no uninsulated wire extends outside of the connectors they are attached to. Stripped wire ends must insert completely into terminal blocks. No uninsulated wiring should be able to come in contact with fingers, tools or clothing during normal operation. 3-4 Teledyne Analytical Instruments Trace Oxygen Analyzer Installation 3 Primary Input Power: The power cord receptacle and fuse block are located in the same assembly. Insert the female plug end of the power cord into the power cord receptacle. CAUTION: Power is applied to the instrument's circuitry as long as the instrument is connected to the power source. The switch on the front panel is for switching power on or off to the displays and outputs only. The universal power supply requires a 85–250 V ac, 47-63 Hz power source. Fuse Installation: The fuse block, at the right of the power cord receptacle, accepts US or European size fuses. A jumper replaces the fuse in whichever fuse receptacle is not used. Fuses are not installed at the factory. Be sure to install the proper fuse as part of installation. (See Fuse Replacement in chapter 5, maintenance.) Analog Outputs: There are four DC output signal connectors with spring terminals on the panel. There are two wires per output with the polarity noted. See Figure 3-4. The outputs are: 0–1 V dc % of Range: Voltage rises linearly with increasing oxygen, from 0 V at 0 ppm to 1 V at full scale ppm. (Full scale = 100% of programmable range.) 0–1 V dc Range ID: 0.25 V = Low Range, 0.5 V = Medium Range, 0.75 V = High Range, 1 V = Air Cal Range. 4–20 mA dc % Range: Current increases linearly with increasing oxygen, from 4 mA at 0 ppm to 20 mA at full scale ppm. (Full scale = 100% of programmable range.) 4–20 mA dc Range ID: 8 mA = Low Range, 12 mA = Medium Range, 16 mA = High Range, 20 mA = Air Cal Range. Figure 3-4: Analog Output Connections Teledyne Analytical Instruments 3-5 3 Installation Model 3000TA Alarm Relays: The three alarm-circuit connectors are spring terminals for making connections to internal alarm relay contacts. Each provides a set of Form C contacts for each type of alarm. Each has both normally open and normally closed contact connections. The contact connections are indicated by diagrams on the rear panel. They are capable of switching up to 3 amperes at 250 V ac into a resistive load. See Figure 3-5. The connectors are: Threshold Alarm 1: • Can be configured as high (actuates when concentration is above threshold), or low (actuates when concentration is below threshold). • Can be configured as failsafe or nonfailsafe. • Can be configured as latching or nonlatching. • Can be configured out (defeated). Threshold Alarm 2: • Can be configured as high (actuates when concentration is above threshold), or low (actuates when concentration is below threshold). • Can be configured as failsafe or nonfailsafe. • Can be configured as latching or nonlatching. • Can be configured out (defeated). System Alarm: Actuates when DC power supplied to circuits is unacceptable in one or more parameters. Permanently configured as failsafe and latching. Cannot be defeated. Actuates if self test fails. (Reset by pressing button to remove power. Then press again and any other button EXCEPT System to resume. Further detail can be found in chapter 4, section 4-5. Figure 3-5: Types of Relay Contacts 3-6 Teledyne Analytical Instruments Trace Oxygen Analyzer Installation 3 Digital Remote Cal Inputs: Accept 0 V (off) or 24 V dc (on) inputs for remote calibration control. (See Remote Calibration Protocol below.) Zero: Floating input. 5 to 24 V input across the + and – terminals puts the analyzer into the Zero mode. Either side may be grounded at the source of the signal. 0 to 1 volt across the terminals allows Zero mode to terminate when done. A synchronous signal must open and close the external zero valve appropriately. See 3.3.3 Remote Probe Connector. (The –C option internal valves operate automatically.) Span: Floating input. 5 to 24 V input across the + and – terminals puts the analyzer into the Span mode. Either side may be grounded at the source of the signal. 0 to 1 volt across the terminals allows Span mode to terminate when done. A synchronous signal must open and close the external span valve appropriately. See 3.3.3 Remote Probe Connector. (The –C option internal valves operate automatically.) Cal Contact: This relay contact is closed while analyzer is spanning and/or zeroing. (See Remote Calibration Protocol below.) Remote Calibration Protocol: To properly time the Digital Remote Cal Inputs to the Model 3000TA Analyzer, the customer's controller must monitor the Cal Relay Contact. When the contact is OPEN, the analyzer is analyzing, the Remote Cal Inputs are being polled, and a zero or span command can be sent. When the contact is CLOSED, the analyzer is already calibrating. It will ignore your request to calibrate, and it will not remember that request. Once a zero or span command is sent, and acknowledged (contact closes), release it. If the command is continued until after the zero or span is complete, the calibration will repeat and the Cal Relay Contact (CRC) will close again. For example: 1) Test the CRC. When the CRC is open, Send a zero command until the CRC closes (The CRC will quickly close.) 2) When the CRC closes, remove the zero command. 3) When CRC opens again, send a span command until the CRC closes. (The CRC will quickly close.) 4) When the CRC closes, remove the span command. When CRC opens again, zero and span are done, and the sample is being analyzed. Teledyne Analytical Instruments 3-7 3 Installation Model 3000TA Note: The Remote Probe connector (paragraph 3.3.3) provides signals to ensure that the zero and span gas valves will be controlled synchronously. If you have the –C Internal valve option—which includes additional zero and span gas inputs— the 3000T automatically regulates the zero, span and sample gas flow. Range ID Relays: Four dedicated Range ID relay contacts. The first three ranges are assigned to relays in ascending order—Low range is assigned to Range 1 ID, Medium range is assigned to Range 2 ID, and High range is assigned to Range 3 ID. The fourth range is reserved for the Air Cal Range (25%). Network I/O: A serial digital input/output for local network protocol. At this printing, this port is not yet functional. It is to be used in future options to the instrument. RS-232 Port: The digital signal output is a standard RS-232 serial communications port used to connect the analyzer to a computer, terminal, or other digital device. It requires a standard 9-pin D connector. The data is status information, in digital form, updated every two seconds. Status is reported in the following order: • • • • • The concentration in ppm or percent The range in use (HI, MED, LO) The span of the range (0-100 ppm, etc) Which alarms—if any—are disabled (AL–x DISABLED) Which alarms—if any—are tripped (AL–x ON). Each status output is followed by a carriage return and line feed. Three input functions using RS-232 have been implemented to date. They are described in Table 3-1. Table 3-1: Commands via RS-232 Input Command Description as Immediately starts an autospan. az Immediately starts an autozero. st Toggling input. Stops/Starts any status message output from the RS-232, until st is sent again. The RS-232 protocol allows some flexibility in its implementation. Table 3-2 lists certain RS-232 values that are required by the 3000TA implementation. 3-8 Teledyne Analytical Instruments Trace Oxygen Analyzer Installation 3 Table 3-2: Required RS-232 Options Parameter Baud Byte Parity Stop Bits Message Interval Setting 2400 8 bits none 1 2 seconds 3.3.3 Remote Probe Connector The 3000TA is a single-chassis instrument, which has no Remote Probe Unit. Instead, the Remote Probe connector is used as another method for controlling external sample/zero/span gas valves. See Figure 3-6. Figure 3-6: Remote Probe Connector Pinouts The voltage from these outputs is nominally 0 V for the OFF and 15 V dc for the ON conditions. The maximum combined current that can be pulled from these output lines is 100 mA. (If two lines are ON at the same time, each must be limited to 50 mA, etc.) If more current and/or a different voltage is required, use a relay, power amplifier, or other matching circuitry to provide the actual driving current. In addition, each individual line has a series FET with a nominal ON resistance of 5 ohms (9 ohms worst case). This could limit the obtainable voltage, depending on the load impedance applied. See Figure 3-7. Teledyne Analytical Instruments 3-9 3 Installation Model 3000TA Figure 3-7: FET Series Resistance 3.4 Installing the Micro-Fuel Cell The Micro-Fuel Cell is not installed in the cell block when the instrument is shipped. Install it before the analyzer is placed in service. Once it is expended, or if the cell is exposed to air for too long, the Micro-Fuel Cell will need to be replaced. The cell could also require replacement if the instrument has been idle for too long. When the micro-Fuel Cell needs to be installed or replaced, follow the procedures in chapter 5, Maintenance, for removing and installing cells. 3.5 Testing the System Before plugging the instrument into the power source: • Check the integrity and accuracy of the gas connections. Make sure there are no leaks. • Check the integrity and accuracy of the electrical connections. Make sure there are no exposed conductors. • Verify that the restriction device has been properly installed (see section 3.3.1). • Check that inlet sample pressure is within the accepted range (see section 3.3.1). Power up the system, and test it by performing the following operations: 1. Repeat the Self-Diagnostic Test as described in chapter 4, section 4.3.5. 3-10 Teledyne Analytical Instruments Trace Oxygen Analyzer Operation 4 Operation 4.1 Introduction Once the analyzer has been installed, it can be configured for your application. To do this you will: • • • • Set system parameters: • Establish a security password, if desired, requiring Operator to log in. • Establish and start an automatic calibration cycle, if desired. Calibrate the instrument. Define the three user selectable analysis ranges. Then choose autoranging or select a fixed range of analysis, as required. Set alarm setpoints, and modes of alarm operation (latching, failsafe, etc). Before you configure your 3000TA these default values are in effect: Ranges: LO = 100 ppm, MED = 1000 ppm, HI = 10,000 ppm. Auto Ranging: ON Alarm Relays: Defeated, 1000 ppm, HI, Not failsafe, Not latching. Zero: Auto, every 0 days at 0 hours. Span: Auto, at 000008.00 ppm, every 0 days at 0 hours. If you choose not to use password protection, the default password is automatically displayed on the password screen when you start up, and you simply press Enter for access to all functions of the analyzer. Teledyne Analytical Instruments 4-1 4 Operation 4.2 Model 3000TA Using the Data Entry and Function Buttons Data Entry Buttons: The < > arrow buttons select options from the menu currently being displayed on the VFD screen. The selected option blinks. When the selected option includes a modifiable item, the Δ∇ arrow buttons can be used to increment or decrement that modifiable item. The Enter button is used to accept any new entries on the VFD screen. The Escape button is used to abort any new entries on the VFD screen that are not yet accepted by use of the Enter button. Figure 4-1 shows the hierarchy of functions available to the operator via the function buttons. The six function buttons on the analyzer are: • Analyze. This is the normal operating mode. The analyzer monitors the oxygen content of the sample, displays the percent of oxygen, and warns of any alarm conditions. • System. The system function consists of six subfunctions that regulate the internal operations of the analyzer: • Auto-Cal setup • Password assignment • Self -Test initiation • Checking software version • Logging out. • Zero. Used to set up a zero calibration. • Span. Used to set up a span calibration. • Alarms. Used to set the alarm setpoints and determine whether each alarm will be active or defeated, HI or LO acting, latching, and/or failsafe. • Range. Used to set up three analysis ranges that can be switched automatically with auto-ranging or used as individual fixed ranges. Any function can be selected at any time by pressing the appropriate button (unless password restrictions apply). The order as presented in this manual is appropriate for an initial setup. Each of these functions is described in greater detail in the following procedures. The VFD screen text that accompanies each operation is reproduced, at the appropriate point in the procedure, in a Monospaced type style. Pushbutton names are printed in Oblique type. 4-2 Teledyne Analytical Instruments Trace Oxygen Analyzer Operation 4 ANALYZE SYSTEM Perform Oxygen Analysis of the Sample SPAN TRAK/HLD ZERO Set Instrument Span Perform Self-Diagnostic Test ALARMS Set Instrument Zero Initiate Automatic Calibration RANGE Set Alarm Setpoints Confrigure Mode of Alarm Operation Set Password Define Analysis Ranges Logout Show Negative Figure 4-1: Hierarchy of Functions and Subfunctions 4.3 The System Function The subfuctions of the System function are described below. Specific procedures for their use follow the descriptions: • • Auto-Cal: Used to define an automatic calibration sequence and/or start an Auto-Cal. PSWD: Security can be established by choosing a 5 digit password (PSWD) from the standard ASCII character set. (See Installing or Changing a Password, below, for a table of ASCII characters available.) Once a unique password is assigned and activated, the operator MUST enter the UNIQUE password to gain access to set-up functions which alter the instrument's operation, such as setting the instrument span or zero setting, adjusting the alarm setpoints, or defining analysis ranges. Teledyne Analytical Instruments 4-3 4 Operation • • • • • • Model 3000TA After a password is assigned, the operator must log out to activate it. Until then, anyone can continue to operate the instrument without entering the new password. Only one password can be defined. Before a unique password is assigned, the system assigns TETAI by default. This allows access to anyone. After a unique password is assigned, to defeat the security, the password must be changed back to TETAI. Logout: Logging out prevents an unauthorized tampering with analyzer settings. More: Select and enter More to get a new screen with additional subfunctions listed. Self–Test: The instrument performs a self-diagnostic test to check the integrity of the power supply, output boards and amplifiers. Version: Displays Manufacturer, Model, and Software Version of instrument. Show Negative: The operator selects whether display can show negative oxygen readings or not. TRAK/HLD: The operator sets whether the instrument analog outputs track the concentration change during calibration and sets a time delay for the concentration alarms after calibration 4.3.1 Tracking the Oxygen Readings during Calibration and Alarm delay The user has the option of setting the preferenc as to whether the analog outputs track the display readings during calibration or not. To set the preference, press the System key once and the first System menu will appear in the VFD display: TRAK/HLD Auto-Cal PSWD Logout More TRAK/HLD should be blinking. To enter this system menu press the Enter key once: Output Sttng: TRACK Alarm Dly: 10 min Or Output Sttng: HOLD Alarm Dly: 10 min 4-4 Teledyne Analytical Instruments Trace Oxygen Analyzer Operation 4 In the first line, TRACK or HOLD should be blinking. The operator can toggle between TRACK and HOLD with the Up or Down keys. When TRACK is selected, the analog outputs (0-1 VDC and 4-20 ma) and the range ID contacts will track the instrument readings during calibration (either zero or span). TRACK is the factory default. When HOLD is selected, the analog outputs (0-1 VDC and 4-20 ma) and the range ID contacts will freeze on their last state before entering one of the calibration modes. When the instrument returns to the Analyze mode, either by a successful or an aborted calibration, there will be a three-minute delay before the analog outputs and the range ID contacts start tracking again. The concentration alarms freeze on their last state before entering calibration regardless of selecting HOLD or TRACK. But, when HOLD is selected the concentration alarms will remain frozen for the time displayed in the second line of the TRAK/HLD menu after the analyzer returns to the Analyze mode. The factory default is three minutes, but the delay time is programmable. To adjust to delay time use the Left or Right arrow keys. When the time displayed on the second line blinks, it can be adjusted by Pressing the Up or Down keys to increase or decrease its value. The minimum delay is 1 minute, the maximum is 30. This preference is stored in non-volatile memory so that it is recovered if power is removed from the instrument. 4.3.2 Setting up an Auto-Cal When proper automatic valving is connected (see chapter 3, installation), the Analyzer can cycle itself through a sequence of steps that automatically zero and span the instrument. Note: If you require highly accurate Auto-Cal timing, use external Auto-Cal control where possible. The internal clock in the Model 3000TA is acurate to 2-3 %. Accordingly, internally scheduled calibrations can vary 2-3 % per day. To setup an Auto–Cal cycle: Choose System from the Function buttons. The LCD will display five subfunctions. TRAK/HLD Auto—Cal PSWD Logout More Teledyne Analytical Instruments 4-5 4 Operation Model 3000TA Use < > arrows to blink Auto—Cal, and press Enter. A new screen for Span/Zero set appears. Span OFF Nxt: 0d 0h Zero OFF Nxt: 0d 0h Press < > arrows to blink Span (or Zero), then press Enter again. (You won’t be able to set OFF to ON if a zero interval is entered.) A Span Every ... (or Zero Every ...) screen appears. Span Every 0 d Start 0 h from now Use Δ∇ arrows to set an interval value, then use < > arrows to move to the start-time value. Use Δ∇ arrows to set a start-time value. To turn ON the Span and/or Zero cycles (to activate Auto-Cal): Press System again, choose Auto—Cal, and press Enter again. When the Span/ Zero values screen appears, use the < > arrows to blink the Span (or Zero) OFF/ON field. Use Δ∇ arrows to set the OFF/ON field to ON. You can now turn these fields ON because there is a nonzero span interval defined. 4.3.3 Password Protection If a password is assigned, then setting the following system parameters can be done only after the password is entered: span and zero settings, alarm setpoints, analysis range definitions, switching between autoranging and manual override, setting up an auto-cal, and assigning a new password. However, the instrument can still be used for analysis or for initiating a selftest without entering the password. If you have decided not to employ password security, use the default password TETAI. This password will be displayed automatically by the microprocessor. The operator just presses the Enter key to be allowed total access to the instrument’s features. NOTE: If you use password security, it is advisable to keep a copy of the password in a separate, safe location. 4.3.3.1 Entering the Password To install a new password or change a previously installed password, you must key in and ENTER the old password first. If the default password is in effect, pressing the ENTER button will enter the default TETAI password for you. 4-6 Teledyne Analytical Instruments Trace Oxygen Analyzer Operation 4 Press System to enter the System mode. TRAK/HLD—Cal PSWD Logout More Use the < > arrow keys to scroll the blinking over to PSWD, and press Enter to select the password function. Either the default TETAI password or AAAAA place holders for an existing password will appear on screen depending on whether or not a password has been previously installed. T E T AI Enter PWD or AAAAA Enter PWD The screen prompts you to enter the current password. If you are not using password protection, press Enter to accept TETAI as the default password. If a password has been previously installed, enter the password using the < > arrow keys to scroll back and forth between letters, and the Δ∇ arrow keys to change the letters to the proper password. Press Enter to enter the password. If the password is accepted, the screen will indicate that the password restrictions have been removed and you have clearance to proceed. PSWD Restrictions Removed In a few seconds, you will be given the opportunity to change this password or keep it and go on. Change Password? =Yes =No Press Escape to move on, or proceed as in Changing the Password, below. 4.3.3.2 Installing or Changing the Password If you want to install a password, or change an existing password, proceed as above in Entering the Password. When you are given the opportunity to change the password: Change Password? =Yes =No Teledyne Analytical Instruments 4-7 4 Operation Model 3000TA Press Enter to change the password (either the default TBEAI or the previously assigned password), or press Escape to keep the existing password and move on. If you chose Enter to change the password, the password assignment screen appears. TE T AI To Proceed or AAAAA To Proceed Enter the password using the < > arrow keys to move back and forth between the existing password letters, and the Δ∇ arrow keys to change the letters to the new password. The full set of 94 characters available for password use are shown in the table below. Characters Available for Password Definition: A K U _ i s } ) 3 = B L V ` j t → * 4 > C M W a k u ! + 5 ? D N X b l v " ' 6 @ E O Y c m w # 7 F P Z d n x $ . 8 G Q [ e o y % / 9 H R ¥ f p z & 0 : I S ] g q { ' 1 ; J T ^ h r | ( 2 < When you have finished typing the new password, press Enter. A verification screen appears. The screen will prompt you to retype your password for verification. AAAAA Retype PWD To Verify Wait a moment for the entry screen. You will be given clearance to proceed. AAAAA TO Proceed Use the arrow keys to retype your password and press Enter when finished. Your password will be stored in the microprocessor and the system 4-8 Teledyne Analytical Instruments Trace Oxygen Analyzer Operation 4 will immediately switch to the Analyze screen, and you now have access to all instrument functions. If all alarms are defeated, the Analyze screen appears as: 0.0 ppm Anlz Range: 0 — 100 If an alarm is tripped, the second line will change to show which alarm it is: 0.0 ppm Anlz AL—1 NOTE: If you log off the system using the logout function in the system menu, you will now be required to re-enter the password to gain access to Span, Zero, Alarm, and Range functions. 4.3.4 Logout The Logout function provides a convenient means of leaving the analyzer in a password protected mode without having to shut the instrument off. By entering Logout, you effectively log off the instrument leaving the system protected against use until the password is reentered. To log out, press the System button to enter the Syste`m function. TRAK/HLD Auto—Cal PSWD Logout More Use the < > arrow keys to position the blinking over the Logout function, and press Enter to Log out. The screen will display the message: Protected Until Password Reentered 4.3.5 System Self-Diagnostic Test The Model 3000TA has a built-in self-diagnostic testing routine. Preprogrammed signals are sent through the power supply, output board and sensor circuit. The return signal is analyzed, and at the end of the test the status of each function is displayed on the screen, either as OK or as a number between 1 and 3. (See System Self Diagnostic Test in chapter 5 for number code.) The self diagnostics are run automatically by the analyzer whenever the instrument is turned on, but the test can also be run by the operator at will. To initiate a self diagnostic test during operation: Teledyne Analytical Instruments 4-9 4 Operation Model 3000TA Press the System button to start the System function. TRAK/HLD Auto—Cal PSWD Logout More Use the < > arrow keys to blink More, then press Enter. Version Self—Test Use the < > arrow keys again to move the blinking to the Self–Test function. The screen will follow the running of the diagnostic. RUNNING DIAGNOSTIC Testing Preamp — 83 During preamp testing there is a countdown in the lower right corner of the screen. When the testing is complete, the results are displayed. Power: OK Analog: OK Preamp: 3 The module is functioning properly if it is followed by OK. A number indicates a problem in a specific area of the instrument. Refer to Chapter 5 Maintenance and Troubleshooting for number-code information. The results screen alternates for a time with: Press Any Key To Continue... Then the analyzer returns to the initial System screen. 4.3.6 Version Screen Move the < > arrow key to More and press Enter. With Version blinking, press Enter. The screen displays the manufacturer, model, and software version information. 4.3.7 Showing Negative Oxygen Readings For software version 1.4.4 or later, the instrument only displays oxygen readings that are positive or zero. The instrument can be reconfigured to show negative readings if sensor output drifts below zero. This situation may arise after the instrument has been zeroed, as time progresses the sensor may drift below the zero calibration setpoint. To show negative oxygen readings on the display: - Press the System key 4-10 Teledyne Analytical Instruments Trace Oxygen Analyzer Operation 4 TRAK/HLD Auto-Cal PSWD Logout More - Use the Right or Left arrow keys and select More. Press Enter. Version Self-Test Show_Negative=NO - Use the Right or Left arrow keys and select “Show_Negative=NO”. - Use the Up or Down key to toggle from NO to YES. - Press the Escape key twice to return to the analyze mode. This preference is stored in non-volatile memory, so this configuration is remembered after a power shutdown. If the instrument is cold started, it will go back to default (not showingg negative oxygen readings). 4.4 The Zero and Span Functions Zeroing is not required in order to achieve the published accuracy specification of this unit. Zeroing will eliminate offset error contributed by sensor, electronics, and internal and external sampling system and improve performance beyond published specification limits. The analyzer is calibrated using zero and span gases. Any suitable oxygen-free gas can be used for zero gas as long as it is known to be oxygen free and does not react adversely with the sample system. Although the instrument can be spanned using air, a span gas with a known oxygen concentration in the range of 70–90% of full scale of the range of interest is recommended. Since the oxygen concentration in air is 20.9% (209,000 ppm), the cell can take a long time to recover if the instrument is used for trace oxygen analysis immediately following calibration in air. Connect the calibration gases to the analyzer according to the instructions given in Section 3.4.1, Gas Connections, observing all the prescribed precautions. Teledyne Analytical Instruments 4-11 4 Operation Model 3000TA Shut off the gas pressure before connecting it to the analyzer, and be sure to limit the pressure to 40 psig or less when turning it back on. Readjust the gas pressure into the analyzer until the flowrate (as read on the analyzer’s SLPM flowmeter) settles between 0.1 and 2.4 SLPM (approximately 0.2 - 5 SCFH). If you are using password protection, you will need to enter your password to gain access to either of these functions. Follow the instructions in sections 4.3.3 to enter your password. Once you have gained clearance to proceed, you can enter the Zero or Span function. 4.4.1 Zero Cal The Zero button on the front panel is used to enter the zero calibration function. Zero calibration can be performed in either the automatic or manual mode. In the automatic mode, an internal algorithm compares consecutive readings from the sensor to determine when the output is within the acceptable range for zero. In the manual mode, the operator determines when the reading is within the acceptable range for zero. Make sure the zero gas is connected to the instrument. If you get a CELL FAILURE message skip to section 4.4.1.3. 4.4.1.1 Auto Mode Zeroing Press Zero to enter the zero function mode. The screen allows you to select whether the zero calibration is to be performed automatically or manually. Use the Δ∇ arrow keys to toggle between AUTO and MAN zero settling. Stop when AUTO appears, blinking, on the display. Zero: Settling: AUTO To Begin Press Enter to begin zeroing. #### PPM Zero Slope=#### ppm/s The beginning zero level is shown in the upper left corner of the display. As the zero reading settles, the screen displays and updates information on Slope (unless the Slope starts within the acceptable zero range and does not need to settle further). Then, and whenever Slope is less than 0.08 for at least 3 minutes, instead of Slope you will see a countdown: 5 Left, 4 Left, and so fourth. These are five steps in the zeroing process that the system must complete, AFTER settling, before it can go back to Analyze. 4-12 Teledyne Analytical Instruments Trace Oxygen Analyzer Operation 4 #### 4 Left=### PPM Zero ppm/s The zeroing process will automatically conclude when the output is within the acceptable range for a good zero. Then the analyzer automatically returns to the Analyze mode. 4.4.1.2 Manual Mode Zeroing Press Zero to enter the Zero function. The screen that appears allows you to select between automatic or manual zero calibration. Use the Δ∇ keys to toggle between AUTO and MAN zero settling. Stop when MAN appears, blinking, on the display. Zero: Settling: Man To Begin Press Enter to begin the zero calibration. After a few seconds the first of five zeroing screens appears. The number in the upper left hand corner is the first-stage zero offset. The microprocessor samples the output at a predetermined rate. It calculates the differences between successive samplings and displays the rate of change as Slope= a value in parts per million per second (ppm/s). #### ppm Zero Slope=#### ppm/s NOTE: It takes several seconds for the true Slope value to display. Wait about 10 seconds. Then, wait until Slope is sufficiently close to zero before pressing Enter to finish zeroing . Generally, you have a good zero when Slope is less than 0.05 ppm/s for about 30 seconds. When Slope is close enough to zero, press Enter. In a few seconds, the screen will update. Once span settling completes, the information is stored in the microprocessor, and the instrument automatically returns to the Analyze mode. 4.4.1.3 Cell Failure Cell failure in the 3000TA is usually associated with inability to zero the instrument down to a satisfactorily low ppm reading. When this occurs, the 3000TA system alarm trips, and the LCD displays a failure message. #.# ppm Anlz Teledyne Analytical Instruments 4-13 4 Operation Model 3000TA CELL FAIL/ ZERO HIGH Before replacing the cell: a. Check your span gas to make sure it is within specifications. b. Check for leaks downstream from the cell, where oxygen may be leaking into the system. If there are no leaks and the span gas is OK, replace the cell as described in chapter 5, Maintenance. 4.4.2 Span Cal The Span button on the front panel is used to span calibrate the analyzer. Span calibration can be performed using the automatic mode, where an internal algorithm compares consecutive readings from the sensor to determine when the output matches the span gas concentration. Span calibration can also be performed in manual mode, where the operator determines when the span concentration reading is acceptable and manually exits the function. 4.4.2.1 Auto Mode Spanning Press Span to enter the span function. The screen that appears allows you to select whether the span calibration is to be performed automatically or manually. Use the Δ∇ arrow keys to toggle between AUTO and MAN span settling. Stop when AUTO appears, blinking, on the display. Span: Settling: AUTO For Next Press Enter to move to the next screen. Calib. Holding time Cal hold: 5 min This menue allows the operator to set the time the analyzer should be held in the span mode, after the readings of the analyzer settle. Five minutes is the default, but it could be adjusted anywhere from 1 to 60 minutes by using the UP or DOWN keys. (A longer Cal holding time will result in a slightly more accurate calibration. Five (5) minutes is appropriate for most users and applications.) Press Enter to move to the next screen. Span Val: 000008.00 Span Mod # 4-14 Teledyne Analytical Instruments Trace Oxygen Analyzer Operation 4 Use the Δ∇ arrow keys to enter the oxygen-concentration mode. Use the < > arrow keys to blink the digit you are going to modify. Use the Δ∇ arrow keys again to change the value of the selected digit. When you have finished typing in the concentration of the span gas you are using (209000.00 if you are using air), press Enter to begin the Span calibration. #### ppm Span Slope=#### ppm/s The beginning span value is shown in the upper left corner of the display. As the span reading settles, the screen displays and updates information on Slope. Spanning automatically ends when the span output corresponds, within tolerance, to the value of the span gas concentration. Then the instrument automatically returns to the analyze mode. 4.4.2.2 Manual Mode Spanning Press Span to start the Span function. The screen that appears allows you to select whether the span calibration is to be performed automatically or manually. Span: Settling:MAN For Next Use the Δ∇ keys to toggle between AUTO and MAN span settling. Stop when MAN appears, blinking, on the display. Press Enter to move to the next screen. Press Enter to move to the next screen. Calib. Holding time Cal hold: 5 min This menue allows the operator to set the time the analyzer should be held in the auto span mode. It does not affect anything in Manual Mode. Just press Enter to continue. Span Val: 000008.00 Span Mod # Press Δ () to permit modification (Mod #) of span value. Use the arrow keys to enter the oxygen concentration of the span gas you are using (209000.00 if you are using air). The < > arrows choose the digit, and the Δ∇ arrows choose the value of the digit. Press Enter to enter the span value into the system and begin the span calibration. Teledyne Analytical Instruments 4-15 4 Operation Model 3000TA Once the span has begun, the microprocessor samples the output at a predetermined rate. It calculates the difference between successive samplings and displays this difference as Slope on the screen. It takes several seconds for the first Slope value to display. Slope indicates rate of change of the Span reading. It is a sensitive indicator of stability. #### Slope=#### % Span ppm/s When the Span value displayed on the screen is sufficiently stable, press Enter. (Generally, when the Span reading changes by 1 % or less of the full scale of the range being calibrated for a period of ten minutes it is sufficiently stable.) Once Enter is pressed, the Span reading changes to the correct value. The instrument then automatically enters the Analyze function. 4.4.3 Span Failure The analyzer checks the output of the cell at the end of the span. If the raw output of the cell is less than 0.5 uA/ppm O2, the span will not be accepted. The analyzer will return to the previous calibration values, trigger the System Alarm, and display in the VFD: Span Failed!! This message will be shown for five seconds and the instrument shall return to the Analyze mode. In the upper right hand corner of the VFD display “FCAL” will be shown. This message flag will help the operator troubleshoot in case calibration was initiated remotely. To reset the alarm and the flag message, the unit must be turned off by cycling the standby key . It will not reset if the next span cycle is correct. A trace cell is unlikely to fail span. As explained before, when the sensor reaches the end of its useful life, the zero offset begins to rise until the analyzer finds the zero unsatisfactory. Nevertheless, feeding the wrong span gas or electronics failure could set this feature off at the end of the span. Consider this before replacing the cell. 4.5 The Alarms Function The Model 3000TA is equipped with 2 fully adjustable concentration alarms and a system failure alarm. Each alarm has a relay with a set of form “C" contacts rated for 3 amperes resistive load at 250 V ac. See Figure in 4-16 Teledyne Analytical Instruments Trace Oxygen Analyzer Operation 4 Chapter 3, Installation and/or the Interconnection Diagram included at the back of this manual for relay terminal connections. The system failure alarm has a fixed configuration described in chapter 3 Installation. The concentration alarms can be configured from the front panel as either high or low alarms by the operator. The alarm modes can be set as latching or non-latching, and either failsafe or non-failsafe, or, they can be defeated altogether. The setpoints for the alarms are also established using this function. Decide how your alarms should be configured. The choice will depend upon your process. Consider the following four points: 1. Which if any of the alarms are to be high alarms and which if any are to be low alarms? Setting an alarm as HIGH triggers the alarm when the oxygen concentration rises above the setpoint. Setting an alarm as LOW triggers the alarm when the oxygen concentration falls below the setpoint. Decide whether you want the alarms to be set as: • Both high (high and high-high) alarms, or • One high and one low alarm, or • Both low (low and low-low) alarms. 2. Are either or both of the alarms to be configured as failsafe? In failsafe mode, the alarm relay de-energizes in an alarm condition. For non-failsafe operation, the relay is energized in an alarm condition. You can set either or both of the concentration alarms to operate in failsafe or non-failsafe mode. 3. Are either of the alarms to be latching? In latching mode, once the alarm or alarms trigger, they will remain in the alarm mode even if process conditions revert back to non-alarm conditions. This mode requires an alarm to be recognized before it can be reset. In the non-latching mode, the alarm status will terminate when process conditions revert to nonalarm conditions. 4. Are either of the alarms to be defeated? The defeat alarm mode is incorporated into the alarm circuit so that maintenance can be performed under conditions which would normally activate the alarms. The defeat function can also be used to reset a latched alarm. (See procedures, below.) Teledyne Analytical Instruments 4-17 4 Operation Model 3000TA If you are using password protection, you will need to enter your password to access the alarm functions. Follow the instructions in section 4.3.3 to enter your password. Once you have clearance to proceed, enter the Alarm function. Press the Alarm button on the front panel to enter the Alarm function. Make sure that AL–1 is blinking. AL—1 AL—2 Choose Alarm Set up alarm 1 by moving the blinking over to AL–1 using the < > arrow keys. Then press Enter to move to the next screen. AL—1 1000 ppm HI Dft—N Fs—N Ltch—N Five parameters can be changed on this screen: • Value of the alarm setpoint, AL–1 #### ppm (oxygen) • Out-of-range direction, HI or LO • Defeated? Dft–Y/N (Yes/No) • Failsafe? Fs–Y/N (Yes/No) • Latching? Ltch–Y/N (Yes/No). • To define the setpoint, use the < > arrow keys to move the blinking over to AL–1 ####. Then use the Δ∇ arrow keys to change the number. Holding down the key speeds up the incrementing or decrementing. (Remember, the setpoint units are ppm O2.) • To set the other parameters use the < > arrow keys to move the blinking over to the desired parameter. Then use the Δ∇ arrow keys to change the parameter. • Once the parameters for alarm 1 have been set, press Alarms again, and repeat this procedure for alarm 2 (AL–2). • To reset a latched alarm, go to Dft– and then press either Δ two times or ∇ two times. (Toggle it to Y and then back to N.) –OR – Go to Ltch– and then press either Δ two times or ∇ two times. (Toggle it to N and back to Y.) 4-18 Teledyne Analytical Instruments Trace Oxygen Analyzer 4.6 Operation 4 The Range Function The Range function allows the operator to program up to three concentration ranges to correlate with the DC analog outputs. If no ranges are defined by the user, the instrument defaults to: Low = 0–100 ppm Med = 0–1,000 ppm High = 0–10,000 ppm. The Model 3000TA is set at the factory to default to autoranging. In this mode, the microprocessor automatically responds to concentration changes by switching ranges for optimum readout sensitivity. If the current range limits are exceeded, the instrument will automatically shift to the next higher range. If the concentration falls to below 85% of full scale of the next lower range, the instrument will switch to that range. A corresponding shift in the DC percent-of-range output, and in the range ID outputs, will be noticed. The autoranging feature can be overridden so that analog output stays on a fixed range regardless of the oxygen concentration detected. If the concentration exceeds the upper limit of the range, the DC output will saturate at 1 V dc (20 mA at the current output). However, the digital readout and the RS-232 output of the concentration are unaffected by the fixed range. They continue to read accurately with full precision. See Front Panel description in Chapter 1. The automatic air calibration range is always 0-25 % and is not programmable. 4.6.1 Setting the Analog Output Ranges To set the ranges, enter the range function mode by pressing the Range button on the front panel. L—100 M—1000 H—10000 Mode—AUTO Use the < > arrow keys to blink the range to be set: low (L), medium (M), or high (H). Use the Δ∇ arrow keys to enter the upper value of the range (all ranges begin at 0 ppm). Repeat for each range you want to set. Press Enter to accept the values and return to Analyze mode. (See note below.) Teledyne Analytical Instruments 4-19 4 Operation Note: Model 3000TA The ranges must be increasing from low to high, for example, if range 1 is set as 0–100 ppm and range 2 is set as 0–1,000 ppm, range 3 cannot be set as 0– 500 ppm since it is lower than range 2. Ranges, alarms, and spans are always set in ppm units (over the entire 0-250,000 ppm range), even though all concentration-data outputs change from ppm units to percent when the concentration is above 10,000 ppm. 4.6.2 Fixed Range Analysis The autoranging mode of the instrument can be overridden, forcing the analyzer DC outputs to stay in a single predetermined range. To switch from autoranging to fixed range analysis, enter the range function by pressing the Range button on the front panel. Use the < > arrow keys to move the blinking over AUTO. Use the Δ∇ arrow keys to switch from AUTO to FX/LO, FX/MED, or FX/HI to set the instrument on the desired fixed range (low, medium, or high). L—100 M—1000 H—10000 Mode—FX/LO or L—100 M—1000 H—10000 Mode—FX/MED or L—100 M—1000 H—10000 Mode—FX/HI Press Escape to re-enter the Analyze mode using the fixed range. NOTE: When performing analysis on a fixed range, if the oxygen concentration rises above the upper limit (or default value) as established by the operator for that particular range, the output saturates at 1 V dc (or 20 mA). However, the digital readout and the RS-232 output continue to read the true value of the oxygen concentration regardless of the analog output range. 4.7 The Analyze Function Normally, all of the functions automatically switch back to the Analyze function when they have completed their assigned operations. Pressing the Escape button in many cases also switches the analyzer back to the Ana- 4-20 Teledyne Analytical Instruments Trace Oxygen Analyzer Operation 4 lyze function. Alternatively, you can press the Analyze button at any time to return to analyzing your sample. 4.8 Signal Output The standard Model 3000TA Trace Oxygen Analyzer is equipped with two 0–1 V dc analog output terminals accessible on the back panel (one concentration and one range ID), and two isolated 4–20 mA dc current outputs (one concentration and one range ID). See Rear Panel in Chapter 3, Installation, for illustration. The signal output for concentration is linear over the currently selected analysis range. For example, if the analyzer is set on range that was defined as 0–100 ppm O2, then the output would be: ppm O2 Voltage Signal Output (V dc) Current Signal Output (mA dc) 0 0.0 4.0 10 0.1 5.6 20 0.2 7.2 30 0.3 8.8 40 0.4 10.4 50 0.5 12.0 60 0.6 13.6 70 0.7 15.2 80 0.8 16.8 90 0.9 18.4 100 1.0 20.0 The analog output signal has a voltage which depends on the oxygen concentration AND the currently activated analysis range. To relate the signal output to the actual concentration, it is necessary to know what range the instrument is currently on, especially when the analyzer is in the autoranging mode. To provide an indication of the range, a second pair of analog output terminals are used. They generate a steady preset voltage (or current when using the current outputs) to represent a particular range. The following table gives the range ID output for each analysis range: Teledyne Analytical Instruments 4-21 4 Operation Model 3000TA Range LO Voltage (V) 0.25 Current (mA) 8 MED 0.50 12 HI 0.75 16 CAL (0-25%) 1.00 20 IMPORTANT: 4-22 In the event of loss of flow through the analyzer, if the vent is vented to a location of high oxygen content, oxygen will back diffuse through the vent line and in most cases quickly saturate the cell with oxygen which can then require a quite long purge down time for the sensor when then exposed to low oxygen concentrations. In the event that flow is to be interrupted into the analyzer, it is suggested that the user do one of the following: 1. Bag the sensor in nitrogen during this time 2. Install a shut off valve on the vent port of the analyzer or somewhere within the users sample system. Teledyne Analytical Instruments Trace Oxygen Analyzer Maintenance 5 Maintenance 5.1 Routine Maintenance Aside from normal cleaning and checking for leaks at the gas connections, routine maintenance is limited to replacing Micro-Fuel cells and fuses, and recalibration. For recalibration, see Section 4.4 Calibration. WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS MANUAL. 5.2 Cell Replacement The L-2 Micro-Fuel Cell is a sealed electrochemical transducer with no electrolyte to change or electrodes to clean. When the cell reaches the end of its useful life, it is replaced. The spent fuel cell should be discarded according to local regulations. This section describes fuel cell care as well as when and how to replace it. 5.2.1 Storing and Handling Replacement Cells To have a replacement cell available when it is needed, TAI recommends that one spare cell be purchased 9-10 months after commissioning the 3000TA, or shortly before the end of the cell's one year warranty period. CAUTION: Do not stockpile cells. The warranty period starts on the day of shipment. The spare cell should be carefully stored in an area that is not subject to large variations in ambient temperature (75 °F nominal) or to rough handling. Teledyne Analytical Instruments 5-1 5 Maintenance Model 3000TA WARNING: THE SENSOR USED IN THE MODEL 3000TA TRACE OXYGEN ANALYZER USES ELECTROLYTES WHICH CONTAIN TOXIC SUBSTANCES, MAINLY LEAD AND POTASSIUM HYDROXIDE, THAT CAN BE HARMFUL IF TOUCHED, SWALLOWED, OR INHALED. AVOID CONTACT WITH ANY FLUID OR POWDER IN OR AROUND THE UNIT. WHAT MAY APPEAR TO BE PLAIN WATER COULD CONTAIN ONE OF THESE TOXIC SUBSTANCES. IN CASE OF EYE CONTACT, IMMEDIATELY FLUSH EYES WITH WATER FOR AT LEAST 15 MINUTES. CALL PHYSICIAN. (SEE APPENDIX, MATERIAL SAFETY DATA SHEET.) CAUTION: Do not disturb the integrity of the cell package until the cell is to actually be used. If the cell package is punctured and air is permitted to enter, the cell will require an excessively long time to reach zero after installation (1-2 weeks!). 5.2.2 When to Replace a Cell The characteristics of the Micro-Fuel Cell show an almost constant output throughout its useful life and then fall off sharply towards zero at the end. Cell failure in the 3000TA is usually characterized inability to zero the instrument down to a satisfactorily low ppm reading. When this occurs, the 3000TA system alarm trips, and the LCD displays a failure message. #.# ppm Anlz CELL FAIL/ ZERO HIGH Before replacing the cell: a. Check your span gas to make sure it is within specifications. b. Check for leaks downstream from the cell, where oxygen may be leaking into the system. If there are no leaks and the span gas is OK, replace the cell. 5.2.3 Removing the Micro-Fuel Cell The Micro-Fuel cell is located inside the stainless steel cell block behind the front panel (see Figure 5-1). To remove an existing cell: 5-2 Teledyne Analytical Instruments Trace Oxygen Analyzer Maintenance 5 1. Remove power to the instrument by unplugging the power cord at the power source. 2. Open the front panel door by pressing the release button on the top right corner of the door all the way in with a narrow gauge tool, such as a small screwdriver, and releasing it. 3. With one hand placed underneath the cell block ready to catch the Micro-Fuel cell , lift up on the stainless steel gate in front of the cell block. This releases the cell and cell holder from the block. The cell and holder will fall out in your hand. Lift Up Cell Block Gate Micro-Fuel Cell Cell Adaptor (For B-2 or A-2 series cell) O-Ring Cell Holder Figure 5-1: Removing the Micro-Fuel Teledyne Analytical Instruments 5-3 5 Maintenance Model 3000TA 5.2.4 Installing a New Micro-Fuel Cell It is important to minimize the amount of time that a Teledyne Trace Oxygen Sensor is exposed to air during the installation process. The quicker the sensor can be installed into the unit, the faster your TAI O2 sensor will recover to low O2 measurement levels. CAUTION: Do not touch the sensing surface of the cell. It is covered with a delicate Teflon membrane that can leak when punctured. The sensor must be replaced if the membrane is damaged. Before installing a new cell, check the O-ring in the base of the cell holder. Replace if worn or damaged. Place the cell on the holder with the screen side facing down. Note: There is a small location hole drilled in the holder. This hole mates with a guide pin on the bottom rear of the cell block. The hole in the cell block holder must align with the guide pin on the cell block. Step 1. Remove power from instrument. Step 2. Remove the old sensor (if installed) from the analyzer. Step 3. Purge the analyzer at approximately 1 SCFH flow rate with N2 (or applicable sample gas with the sensor holder removed). Step 4. Remove sensor from double bag storage. Step 5. Remove sensor shorting button. Step 6. Place sensor on sensor holder so that the gold contact plate of the sensor is facing up towards the sky. Step 7. Install sensor and sensor holder into cell block. Step 8. With O-ring in place, align the guide pin with the hole on the cell holder. Then, with the holder, lift cell into the cell block. Step 9. Push the gate on the cell block down so that the slots on the side of the gate engage the locating screws on the side of the block. This forces the holder into position and forms a gastight seal. Step 10. Purge system with sample or zero gas. Step 11. Power-up. If steps 4 through 10 are accomplished quickly (elapsed time less than 15 seconds), recovery to less than 1ppm level should occur in less than 8 hours. 5-4 Teledyne Analytical Instruments Trace Oxygen Analyzer Maintenance 5 5.2.5 Cell Warranty The Class L-2 Micro-Fuel cell is used in the Model 3000TA. This cell is a long life cell and is warranted for 1 year from the date of shipment. Note any Addenda attached to the front of this manual for special information applying to your instrument. With regard to spare cells, warranty period begins on the date of shipment. The customer should purchase only one spare cell (per section 5.2.1). Do not attempt to stockpile spare cells. The L-2 cell is not designed applications where CO2 is a major component in the sample, however concentrations of 1,000 ppm or less will not adversely effect the cell performance. Consult TAI for available options for either intermittent or continuous CO2 exposure. If a cell was working satisfactorily, but ceases to function before the warranty period expires, the customer will receive credit toward the purchase of a new cell. If you have a warranty claim, you must return the cell in question to the factory for evaluation. If it is determined that failure is due to faulty workmanship or material, the cell will be replaced at no cost to you. Note: Evidence of damage due to tampering or mishandling will render the cell warranty null and void. 5.3 Fuse Replacement 1. Place small screwdriver in notch, and pry cover off, as shown in Figure 5-2. Figure 5-2: Removing Fuse Block from Housing Teledyne Analytical Instruments 5-5 5 Maintenance Model 3000TA 2. To change between American and European fuses, remove the single retaining screw, flip Fuse Block over 180 degrees, and replace screw. 3. Replace fuse as shown in Figure 5-3. 4. Reassemble Housing as shown in Figure 5-2. American Fuses European Fuses Figure 5-3: Installing Fuses 5.4 System Self Diagnostic Test 1. Press the System button to enter the system mode. 2. Use the < > arrow keys to move to More, and press Enter. 3. Use the < > arrow keys to move to Self-Test, and press Enter. The following failure codes apply: Table 5-1: Self Test Failure Codes Power 0 1 2 3 OK 5 V Failure 15 V Failure Both Failed Analog 0 1 2 3 OK DAC A (0–1 V Concentration) DAC B (0–1 V Range ID) Both Failed Preamp 0 1 2 3 5-6 OK Zero too high Amplifier output doesn't match test input Both Failed Teledyne Analytical Instruments Trace Oxygen Analyzer 5.5 Maintenance 5 Major Internal Components The Micro-Fuel cell is accessed by unlatching and swinging open the front panel, as described earlier. Other internal components are accessed by removing the rear panel and sliding out the entire chassis. See Figure 5-4, below. The gas piping is illustrated in Figure 2-4, and the major electronic components locations are shown in Figure 2-5, in chapter 2. WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS MANUAL. The 3000TA contains the following major components: • Analysis Section Micro Fuel Cell (L-2) Stainless steel cell block Sample system • Power Supply • Microprocessor • Displays 5 digit LED meter 2 line, 20 character, alphanumeric, VFD display • RS-232 Communications Port. See the drawings in the Drawings section in back of this manual for details. x x x x x x x x Figure 5-4: Rear-Panel Screws To detach the rear panel, remove only the eight screws marked with an X. Teledyne Analytical Instruments 5-7 5 Maintenance 5.6 Model 3000TA Cleaning If instrument is unmounted at time of cleaning, disconnect the instrument from the power source. Close and latch the front-panel access door. Clean outside surfaces with a soft cloth dampened slightly with plain clean water. Do not use any harsh solvents such as paint thinner or benzene. For panel-mounted instruments, clean the front panel as prescribed in the above paragraph. DO NOT wipe front panel while the instrument is controlling your process. 5.7 Troubleshooting Problem: Erratic readings of the Oxygen concentration as reported by the analyzer. Possible Cause: The analyzer may have been calibrated in an inaccurate fashion. Solution: Turn the analyzer off, then back on again. Press the System key when prompted by the analyzer "Press System for default Values". This will return the analyzer to its default settings in calibration and zero values. If erratic behavior continues replace the sensor. Possible Cause: Atmospheric Oxygen may be diffusing in through the vent and affecting the oxygen level which the sensor sees. Solution: Increase flow rate and/or length or vent tubing in order to dilute of minimize the diffusion of oxygen from the vent back to the sensor. Problem: Inaccurate zero operation (i.e. the user has zeroed the analyzer accidentally on gas much higher than one would normally use for a zero gas). Solution: Turn the analyzer off, then back on again. Press the System key when prompted by the analyzer "Press System for default Values". This will return the analyzer to its default settings in calibration and zero values. Now proceed to carefully calibrate and zero the analyzer. 5-8 Teledyne Analytical Instruments Trace Oxygen Analyzer Appendix Appendix A-1 Specifications Packaging: General Purpose • Flush panel mount (Standard). • Relay rack mount. Contains either one or two instruments in one 19" relay rack mountable plate (Optional). Sensor: Teledyne L-2 trace analysis Micro-Fuel Cell. Cell Block: 316 stainless steel. Sample System: All wetted parts of 316 stailess steel. 90 % Response Time: 65 seconds at 25 °C (77 °F). Ranges: Three user definable ranges from 0–10 ppm to 0–250,000 ppm, plus air calibration range of 0250,000 ppm (25 %). Autoranging with range ID output. Alarms: One system-failure alarm contact to detect power failure or sensor-zero failure. Two adjustable concentration threshold alarm contacts with fully programmable setpoints. Displays: 2-line by 20-character, VFD screen, and one 5 digit LED display. Digital Interface: Full duplex RS-232 communications port. Power: Universal power supply 85-250 V ac, at 47-63 Hz. Operating Temperature: 0-50 °C (32-122 °F) Teledyne Analytical Instruments A-1 Appendix Model 3000TA Accuracy: ±2% of full scale at constant temperature. ±5% of full scale over operating temperature range, except 0-10 ppm analysis range, once thermal equilibrium is reached. ±1 ppm on 0-10 ppm analysis range, once thermal equilibrium is reached. Analog outputs: 0-1 V dc percent-of-range, 0-1 V dc range ID. 4-20 mA dc—isolated—percent-of-range, 4-20 mA dc—isolated—range ID. Dimensions: 19 cm high, 24.9 cm wide, 31 cm deep (6.96 in high, 8.7 in wide, 12.2 in deep). A-2 Teledyne Analytical Instruments Trace Oxygen Analyzer Appendix A-2 Recommended 2-Year Spare Parts List QTY. PART NUMBER DESCRIPTION 1 C62374 Back Panel Board 1 C62371-B Front Panel Board 1 C62368-A Trace Preamplifier Board 1 C73870-A Main Computer Board 3 F9 Fuse, 1A, 250V 3AG Slow Blow 3 F1275 Fuse, 1A, 250V 5x20mm (European) Slow Blow 1 R1460 Molex Connector for Remote Probe 1 T976 Molex Crimp Terminals for Remote Probe Connector 1 O165 O-ring 1 C6689-L2 Micro-Fuel Cell 1 A68729 Restrictor Kit A minimum charge is applicable to spare parts orders. Note: Orders for replacement parts should include the part number (if available) and the model and serial number of the instrument for which the parts are intended. Orders should be sent to: Teledyne Analytical Instruments 16830 Chestnut Street City of Industry, CA 91749-1580 Phone (626) 934-1500, Fax (626) 934-1500 TWX (910) 584-1887 TDYANYL COID Web: www.teledyne-ai.com or your local representative. Teledyne Analytical Instruments A-3 Appendix Model 3000TA A-3 Drawing List D-62928 Final Assembly/Outline Drawing A-4 19-inch Relay Rack Panel Mount Figure A-1: Single and Dual 19" Rack Mounts NOTE: A-4 The MSDS on this material is available upon request through the Teledyne Environmental, Health and Safety Coordinator. Contact at (626) 934-1592 Teledyne Analytical Instruments Trace Oxygen Analyzer Appendix A-5 3000 SERIES ANALYZERS APPLICATION NOTES ON RESTRICTORS, PRESSURES, AND FLOW RECOMMENDATIONS 3000 series analyzers require reasonably regulated sample pressures. While the 3000 analyzers are not sensitive to variations of incoming pressure (provided they are properly vented to atmospheric pressure) the pressure must be maintained as to provide a useable flow rate trough the analyzer. Any line attached to sample vent should be 1/4 or larger in diameter. FLOW RATE RECOMMENDATIONS: A usable flow rate for a 3000 series analyzer is one which can be measured on the flowmeter. This is basically .2 - 2.4 SLPM . The optimum flow rate is 1 SLPM (mid scale). Note: response time is dependent on flow rate, a low flow rate will result in slow response to O2 changes in the sample stream. The span flow rate should be the approximately same as the sample flow rate. CELL PRESSURE CONCERNS: The sensors used in 3000 series analyzers are optimized to function at atmospheric pressure. At pressures other than atmospheric the diffusion rate of O2 will be different than optimum value. Higher pressures will produce faster O2 diffusion rates resulting in higher O2 reading and shorter cell life. To use a 3000 series analyzer at a cell pressure other than atmospheric, the analyzer must be calibrated with a known calibration gas at the new cell pressure to adjust for the different diffusion rate. Cell pressures below 2/3 atmospheric are not recommended because as they tend to cause excessive internal expansion which may result in seal failure. For operation at cell pressures other than atmospheric care must be taken not to change the sample pressure rapidly or cell damage may occur. For cell pressures above atmospheric, caution must be exercised to avoid over pressuring the cell holder. ( percent analyzers will require some type of cell retainer to prevent the cell from being pushed out by the pressure .) For operation at pressures below atmospheric pressure a suffix C ( clamped) cell is required. RESTRICTION DEVICES: For proper operation, all 3000 series analyzers require a flow restriction device. This device is typically a restrictor or a valve. This restriction device serves two functions in the sample path. The first function is to limit the flow rate of the sample through the analyzer. A restrictor is chosen to operate over a range of pressures and provide a useable flow rate over that range. Teledyne Analytical Instruments A-5 Appendix Model 3000TA The second function that the restriction device provides is a pressure drop. This device is selected to provide the only significant pressure drop in the sample path. RESTRICTOR KIT The current revision of the 3000 series analyzers are supplied with a kit containing two restrictors and a union which are user installed. These parts supplied to give the end user more flexibility when installing the analyzer. The restrictor kit is suitable for high and low positive pressure applications as well as vacuum service ( atmospheric pressure sample) applications ( see manual for installation instructions). The standard restrictor ( BLUE DOT ) is recommended for pressures between 5 PSIG and 50 PSIG. For positive low pressure application ( 5 psig or less ) the un-marked restrictor is better suited . For none pressurized sample applications the marked restrictor should be used and configured for vacuum service. Note: for extremely low positive pressure applications ( less then 2 psig) the vacuum service configuration should provide higher performance ( higher flow rates). For vacuum service the end user must supply a vacuum pump and a by-pass valve for the pump. A vacuum level of 5 10 inches of mercury should provide the optimum flow rate. CAUTION: flow restrictors have very small orifices and may be plugged by small particles ( .005” dia or larger) A sample filter must be included in the sample line prior to the restrictor! ( a 60 micron filter is recommended) 3000TA EXAMPLES: Example 1, with a incoming pressure of 10 psig the std restrictor (blue dot) will provide a flow rate of .76 SLPM. Up-stream of the restrictor the sample line pressure will be 10 psig, while down stream ( including the cell) the pressure will be at atmospheric pressure.( analyzer vented to atmospheric pressure) Note, all other pressure drops in the sample path are insignificant at these flow rates. This insures that the cell operates at atmospheric pressure. At very high flow rates ( off scale of flow-meter), pressure drops other than the restriction device could become significant , and result in pressurizing the cell. Example 2, A 3000TA is configured for vacuum service as follows. The un-marked restrictor is placed in the sample vent port. The down stream end of the restrictor is then connected to a vacuum pump and by-pass valve. The bypass valve is adjusted to provide a flow rate of 1 SLPM. The sample pressure between the pump and the restrictor will be approximately -7 inches of mercury, while the pressure in the balance of the sample system including the cell will be approximately at atmospheric pressure. ( provided the sample flow into the analyzer is not blocked.) BY-PASS: To improve the system response, a by-pass can be added to increase the sample flow rate to the analyzer by a factor of ten. A by-pass provides a sample flow path around the analyzer of 2 - 18 SCFH. typically. A-6 Teledyne Analytical Instruments Trace Oxygen Analyzer Appendix CALIBRATION GAS: 3000 series analyzer requirements for units with Auto-Cal options. The customer must supply a control valves (or restrictors) for any SPAN or ZERO gas source which is attached to the Auto-Cal ports. The valve should be adjusted to the same flow rate as the sample gas . When restrictors are used, the gas pressure must be adjusted to achieve the proper flow rate. OPERATION WITHOUT A RESTRICTOR DEVICE: Operation without a restrictor device is not recommend as mentioned above. A 3000TA without any flow restrictor device was tested on 11-19-97. This results in a flow rate of 2.4 SLPM @ 1 PSIG. This is a cv of 0.023 for the standard sample sys. REFERENCE: FLOW_1.XLS & FLOW_2.XLS for information on flow rates at various pressures. TAI PART NUMBERS RESTRICTOR KIT: UNION (SS) LP. RESTRICTOR STD.. RESTRICTOR NUT FERRULE FERRULE A68729 U11 R2323 ( LOW PRESSURE / VAC. SERVICE ) R2324 BLUE DOT N73 F73 F74 BOTH FERRULES ARE REQUIRED CONVERSIONS: 1 PSI 1 SCFH = = 2.04 INCHES OF MERCURY (in. Hg.) 0.476 SLPM Teledyne Analytical Instruments A-7 Appendix A-8 Model 3000TA Teledyne Analytical Instruments Trace Oxygen Analyzer Appendix A-6 Material Safety Data Sheet Section I – Product Identification Product Name: Micro-Fuel Cells and Super Cells, all classes except A-2C, A-3, and A-5. Electrochemical Oxygen Sensors, all classes except R-19. Mini-Micro-Fuel Cells, all classes. Manufacturer: Address: Phone: Technical Support: Environment, Health and Safety: Date Prepared : Teledyne Electronic Technologies/Analytical Instruments 16830 Chestnut Street, City of Industry, CA 91749 (626) 934-1500 (626) 934-1673 (626) 934-1592 09/17/98 Section II – Hazardous Ingredients/Composition Material or Component TLV C.A.S. # Quantity OSHA PEL ACGIH Lead (Pb) 7439-92-1 3–20 gms 0.05 mg/m3 0.15 mg/m3 Potassium Hydroxide Solution 15% (KOH) 1310-58-3 1–5 ml None 2 mg/m3 Section III – Physical/Chemical Characteristics Material Specific Vapor Appearance Boiling or Compo- Point (°C) Gravity Pressure nent 11.34 1744 na Lead 1320 Potassium Hydroxide Odor Melting Density Evap. Solubility Point Rate in Water (°C) Insoluble Solid, silver na na gray, odorle 328 2.04 na na 360 Teledyne Analytical Instruments na Complete White or slightly yellow, no odor A-9 Appendix Model 3000TA Section IV – Fire and Explosion Hazard Data Flash Point: na Flammable Limits: na LEL: na UEL: na Extinguishing Media: Use extinguishing media appropriate to surrounding fire conditions. No specific agents recommended. Special Fire Fighting Equipment: Wear NIOSH/OSHA approved self-contained breathing apparatus and protective clothing to prevent contact with skin and eyes. Unusual Fire and Explosion Hazards: Emits toxic fumes under fire conditions. Section V – Reactivity Data Stability: Stable Incompatibilities: Aluminum, organic materials, acid chlorides, acid anhydrides, magnesium, copper. Avoid contact with acids and hydrogen peroxide > 52%. Hazardous Decomposition of Byproducts: Hazardous Polymerization: Toxic fumes Will not occur. Conditions to Avoid: Section VI – Health Hazard Data Routes of Entry: Inhalation: Highly unlikely Ingestion: May be fatal if swallowed. Skin: The electrolyte (potassium hydroxide) is corrosive; skin contact may cause irritation or chemical burns. Eyes: The electrolyte (potassium hydroxide) is corrosive; eye contact may cause irritation or severe chemical burns. Acute Effects: The electrolyte is harmful if swallowed, inhaled or adsorbed through the skin. It is extremely destructive to tissue of the mucous membranes, stomach, mouth, upper respiratory tract, eyes and skin. Chronic Effects: Prolonged exposure with the electrolyte has a destructive effect on tissue. Chronic exposure to lead may cause disease of the blood and blood forming organs, kidneys and liver, damage to the reproductive systems and decrease in fertility in men and women, and damage to the fetus of a pregnant woman. Chronic exposure from the lead contained in this product is extremely unlikely. A-10 Teledyne Analytical Instruments Trace Oxygen Analyzer Appendix Signs and Symptoms of Exposure: Contact of electrolyte with skin or eyes will cause a burning sensation and/or feel soapy or slippery to touch. Other symptoms of exposure to lead include loss of sleep, loss of appetite, metallic taste and fatigue. Carcinogenicity: Lead is classified by the IARC as a class 2B carcinogen (possibly carcinogenic to humans) OSHA: Where airborne lead exposures exceed the OSHA action level, refer to OSHA Lead Standard 1910.1025. NTP: na Medical Conditions Generally Aggravated by Exposure: Lead exposure may aggravate disease of the blood and blood forming organs, hypertension, kidneys, nervous and possibly reproductive systems. Those with preexisting skin disorders or eye problems may be more susceptible to the effects of the electrolyte. Emergency First Aid Procedures: In case of contact with the skin or eyes, immediately flush with plenty of water for at least 15 minutes and remove all contaminated clothing. Get medical attention immediately. If ingested, give large amounts of water and DO NOT INDUCE VOMITING. Obtain medical attention immediately. If inhaled, remove to fresh air and obtain medical attention immediately. Section VII – Precautions for Safe Handling and Use NOTE: The oxygen sensors are sealed, and under normal circumstances, the contents of the sensors do not present a health hazard. The following information is given as a guide in the event that a cell leaks. Protective measures during cell replacement: Before opening the bag containing the sensor cell, check the sensor cell for leakage. If the sensor cell leaks, do not open the bag. If there is liquid around the cell while in the instrument, wear eye and hand protection. Cleanup Procedures: Wipe down the area several times with a wet paper towel. Use a fresh towel each time. Contaminated paper towels are considered hazardous waste. Teledyne Analytical Instruments A-11 Appendix Model 3000TA Section VIII – Control Measures Eye Protection: Chemical splash goggles Hand Protection: Rubber gloves Other Protective Clothing: Apron, face shield Ventilation: na Section IX – Disposal Both lead and potassium hydroxide are considered poisonous substances and are regulated under TSCA and SARA Title III. EPA Waste Number: D008 California Waste Number: 181 DOT Information: RQ Hazardous Waste Solid N.O.S. (Lead) Class 9 NA3077 PG III Follow all Federal, State and Local regulations. Section X – References Material Safety Data Sheets from J.T. Baker Chemical, Aldrich, Malinckrodt, ASARCO U.S. Department of Labor form OMB No. 1218-0072 Title 8 California Code of Regulations TSCA SARA Title III CFR 49 CFR 29 CFR 40 NOTE: The above information is believed to be correct and is offered for your information, consideration, and investigation. It should be used as a guide. Teledyne Brown Engineering Analytical Instruments shall not be held liable for any damage resulting from handling or from contact with the above product. A-12 Teledyne Analytical Instruments Trace Oxygen Analyzer Appendix Teledyne Analytical Instruments A-13