Endress + Hauser Operating Instructions Oxymax COS61D Dissolved oxygen sensor With Memosens protocol
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
BA00460C/07/EN/14.16
71341281
Products
Solutions
Operating Instructions
Oxymax COS61D
Dissolved oxygen sensor
With Memosens protocol
Services
Table of contents
Table of contents
1
Document information . . . . . . . . . . . . . . 3
10
1.1
1.2
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2
Basic safety instructions . . . . . . . . . . . . 4
2.1
2.2
2.3
2.4
2.5
Requirements for personnel . . . . . . . . . . . . . . .
Designated use . . . . . . . . . . . . . . . . . . . . . . . .
Occupational safety . . . . . . . . . . . . . . . . . . . . .
Operational safety . . . . . . . . . . . . . . . . . . . . . .
Product safety . . . . . . . . . . . . . . . . . . . . . . . . .
10.1 Maintenance schedule . . . . . . . . . . . . . . . . . .
10.2 Maintenance tasks . . . . . . . . . . . . . . . . . . . . .
10.3 Clean exterior of sensor. . . . . . . . . . . . . . . . . .
10.4 Cleaning the sensor optics . . . . . . . . . . . . . . .
10.5 Wear parts and consumables . . . . . . . . . . . . .
3
Device description, function . . . . . . . . . 6
3.1
3.2
3.3
3.4
Optical measuring principle . . . . . . . . . . . . . . .
Sensor design . . . . . . . . . . . . . . . . . . . . . . . . .
Memosens technology . . . . . . . . . . . . . . . . . . .
Fluorescence cap . . . . . . . . . . . . . . . . . . . . . . .
4
4
4
4
4
5
Maintenance . . . . . . . . . . . . . . . . . . . . . . 25
25
25
25
26
26
11
Accessories . . . . . . . . . . . . . . . . . . . . . . . 28
11.1
11.2
11.3
11.4
11.5
11.6
11.7
Assemblies (selection) . . . . . . . . . . . . . . . . . .
Assembly holder . . . . . . . . . . . . . . . . . . . . . .
Measuring cable . . . . . . . . . . . . . . . . . . . . . .
Zero-point gel . . . . . . . . . . . . . . . . . . . . . . . .
Protection guard . . . . . . . . . . . . . . . . . . . . . .
Cleaning unit . . . . . . . . . . . . . . . . . . . . . . . . .
Calibration vessel . . . . . . . . . . . . . . . . . . . . . .
12
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Incoming acceptance and product
identification . . . . . . . . . . . . . . . . . . . . . . . 9
12.1
12.2
12.3
Spare parts and consumables . . . . . . . . . . . . . 30
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
4.1
4.2
4.3
Incoming acceptance . . . . . . . . . . . . . . . . . . . . 9
Product identification . . . . . . . . . . . . . . . . . . . . 9
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . 10
13
Technical data . . . . . . . . . . . . . . . . . . . . 31
5
Installation . . . . . . . . . . . . . . . . . . . . . . . 11
5.1
5.2
5.3
5.4
Installation conditions . . . . . . . . . . . . . . . . . .
Mounting the sensor . . . . . . . . . . . . . . . . . . .
Installation examples . . . . . . . . . . . . . . . . . . .
Post-installation check . . . . . . . . . . . . . . . . . .
6
Electrical connection . . . . . . . . . . . . . . 19
6.1
6.2
6.3
Connecting the sensor . . . . . . . . . . . . . . . . . . 19
Ensuring the degree of protection . . . . . . . . . . 19
Post-connection check . . . . . . . . . . . . . . . . . . 19
7
Calibration and adjustment . . . . . . . . 20
7.1
7.2
7.3
7.4
Types of calibration . . . . . . . . . . . . . . . . . . . .
Calibration intervals . . . . . . . . . . . . . . . . . . . .
Calibration in air . . . . . . . . . . . . . . . . . . . . . .
Calculation example for the calibration
value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8
Commissioning . . . . . . . . . . . . . . . . . . . . 23
8.1
8.2
8.3
Function check . . . . . . . . . . . . . . . . . . . . . . .
Sensor calibration . . . . . . . . . . . . . . . . . . . . .
Cleaning the sensor automatically . . . . . . . . .
9
Troubleshooting . . . . . . . . . . . . . . . . . . 24
9.1
9.2
Trouble shooting instructions . . . . . . . . . . . . . 24
Check sensor . . . . . . . . . . . . . . . . . . . . . . . . . 24
2
6
7
8
8
28
28
28
28
29
29
29
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
11
12
15
18
20
20
20
21
23
23
23
Endress+Hauser
Oxymax COS61D
Document information
1
Document information
1.1
Warnings
Structure of information
LDANGER
Causes (/consequences)
Consequences of non-compliance
(if applicable)
‣ Corrective action
LWARNING
Causes (/consequences)
Consequences of non-compliance
(if applicable)
‣ Corrective action
LCAUTION
Causes (/consequences)
Consequences of non-compliance
(if applicable)
‣ Corrective action
NOTICE
Cause/situation
Consequences of non-compliance
(if applicable)
‣ Action/note
1.2
Symbol
Meaning
This symbol alerts you to a dangerous situation.
Failure to avoid the dangerous situation will result in a fatal or serious
injury.
This symbol alerts you to a dangerous situation.
Failure to avoid the dangerous situation can result in a fatal or serious
injury.
This symbol alerts you to a dangerous situation.
Failure to avoid this situation can result in minor or more serious injuries.
This symbol alerts you to situations which may result in damage to
property.
Symbols
Meaning
Additional information, tips
Permitted or recommended
Not permitted or not recommended
Reference to device documentation
Reference to page
Reference to graphic
Endress+Hauser
Result of a step
3
Basic safety instructions
Oxymax COS61D
2
Basic safety instructions
2.1
Requirements for personnel
• Installation, commissioning, operation and maintenance of the measuring system may
be carried out only by specially trained technical personnel.
• The technical personnel must be authorized by the plant operator to carry out the
specified activities.
• The electrical connection may be performed only by an electrical technician.
• The technical personnel must have read and understood these Operating Instructions
and must follow the instructions contained therein.
• Measuring point faults may be repaired only by authorized and specially trained
personnel.
Repairs not described in the Operating Instructions provided may only be carried out
directly by the manufacturer or by the service organization.
2.2
Designated use
The oxygen sensor is suitable for continuous measurement of dissolved oxygen in water.
The main areas of application are:
• Wastewater treatment plants
– Oxygen measurement and regulation in the activated sludge basin for a highly
efficient biological cleaning process
– Monitoring the oxygen content in the wastewater treatment plant outlet
• Water monitoring
Oxygen measurement in rivers, lakes or seas as an indicator of the water quality
• Water treatment
Oxygen measurement for status monitoring, e.g. of drinking water (oxygen enrichment,
corrosion protection etc.)
• Fish farming
Oxygen measurement and regulation for optimum living and growth conditions
Use of the device for any purpose other than that described, poses a threat to the safety of
people and of the entire measuring system and is therefore not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.
2.3
Occupational safety
As the user, you are responsible for complying with the following safety conditions:
• Installation guidelines
• Local standards and regulations
Electromagnetic compatibility
• The product has been tested for electromagnetic compatibility in accordance with the
applicable European standards for industrial applications.
• The electromagnetic compatibility indicated applies only to a product that has been
connected in accordance with these Operating Instructions.
2.4
4
Operational safety
1.
Before commissioning the entire measuring point, verify that all connections are
correct. Ensure that electrical cables and hose connections are undamaged.
2.
Do not operate damaged products, and safeguard them to ensure that they are not
operated inadvertently. Label the damaged product as defective.
Endress+Hauser
Oxymax COS61D
Basic safety instructions
3.
If faults cannot be rectified:
Take the products out of operation and safeguard them to ensure that they are not
operated inadvertently.
LCAUTION
Cleaning not switched off during calibration or maintenance activities
Risk of injury due to medium or cleaning agent
‣ If a cleaning system is connected, switch if off before removing a sensor from the
medium.
‣ If you wish to check the cleaning function and have therefore not switched off the
cleaning system, please wear protective clothing, goggles and gloves or take other
appropriate measures.
2.5
Product safety
The product is designed to meet state-of-the-art safety requirements, has been tested, and
left the factory in a condition in which it is safe to operate. The relevant regulations and
European standards have been observed.
Endress+Hauser
5
Device description, function
Oxymax COS61D
3
Device description, function
3.1
Optical measuring principle
Sensor structure
Oxygen-sensitive molecules (markers) are integrated into an optically active layer
(fluorescence layer).
The surface of the fluorescence layer is in contact with the medium.
The sensor optics are directed at the underside of the fluorescence layer.
Measurement process (principle of quenching)
If the sensor is immersed in the medium, an equilibrium is very quickly established
between the oxygen partial pressure in both the medium and the fluorescence layer.
1.
The sensor optics send green light pulses to the fluorescence layer.
2.
The markers "answer" (fluoresce) with red light pulses.
The duration and intensity of the response signals is directly dependent on the
oxygen contents and the partial pressure.
If the medium is free from oxygen, the response signals are long and very intense.
Oxygen molecules mask the marker molecules. As a result, the response signals are
shorter and less intense.
Measurement result
‣ The sensor returns a signal that is in proportion to the oxygen concentration in the
medium.
The medium temperature and air pressure are already taken into account calculated in the
sensor.
In addition to the standard values of concentration, saturation index and partial pressure,
the sensor also returns a raw measured value in µs. The value corresponds to the decay
time of the fluorescence and is approx. 20 μs in air and approx. 60 μs in oxygen-free
media.
To ensure an optimum measurement result, the current air pressure should be
entered at the transmitter during calibration. In the case of saline media, you should
also enter the salinity value. For measurement in the units %Vol or %SAT, you must
also enter the current operating pressure in measuring mode.
6
Endress+Hauser
Oxymax COS61D
Device description, function
3.2
Sensor design
1
5
2
6
3
7
4
8
A0013516
1
Sensor structure
1
Sensor cable
5
Threaded connection
2
Sensor shaft
6
Detector
3
O-ring
7
Emitter diode
4
Protection guard
8
Fluorescence cap
The sensor consists of the following function units:
• Sensor shaft
• Sensor head with optics (emitter and detector)
• Fluorescence cap
• Protection guard
As an alternative to the protection guard, you can use a COR3 spray head (→ 29) for
immersion operation with cleaning.
Endress+Hauser
7
Device description, function
Oxymax COS61D
3.3
Memosens technology
Sensors with Memosens protocol have an integrated electronics unit that stores calibration
data and other information. Once the sensor has been connected, the sensor data are
transferred automatically to the transmitter and used to calculate the measured value. You
can call up the sensor data via the relevant DIAG menu.
Digital sensors can store measuring system data in the sensor. These include the following:
• Manufacturer data
– Serial number
– Order code
– Date of manufacture
• Calibration data
– Calibration date
– Calibration values
– Number of calibrations
– Serial number of the transmitter used to perform the last calibration
• Operating data
– Temperature application range
– Date of initial commissioning
– Hours of operation under extreme conditions
3.4
Fluorescence cap
The oxygen dissolved in the medium is diffused into the fluorescencecap. Suitable flow is
not necessarily mandatory but it does improve the speed at which the measuring system
responds and ensures a more representative measured value compared to a measurement
in static medium.
The cap is only permeable for dissolved gases. Other substances dissolved in the liquid
phase e.g. ionic substances, will not penetrate through the membrane. Therefore, medium
conductivity has no impact on the measuring signal.
8
Endress+Hauser
Oxymax COS61D
Incoming acceptance and product identification
4
Incoming acceptance and product
identification
4.1
Incoming acceptance
1.
Verify that the packaging is undamaged.
Notify your supplier of any damage to the packaging.
Keep the damaged packaging until the matter has been settled.
2.
Verify that the contents are undamaged.
Notify your supplier of any damage to the delivery contents.
Keep the damaged products until the matter has been settled.
3.
Check the delivery for completeness.
Check it against the delivery papers and your order.
4.
Pack the product for storage and transportation in such a way that it is protected
against impact and moisture.
The original packaging offers the best protection.
The permitted ambient conditions must be observed (see "Technical data").
If you have any questions, please contact your supplier or your local sales center.
4.2
Product identification
4.2.1
Nameplate
The nameplate provides you with the following information on your device:
• Manufacturer identification
• Order code
• Extended order code
• Serial number
• Safety information and warnings
‣ Compare the data on the nameplate with your order.
4.2.2
Product identification
Product page
www.endress.com/cos61d
Interpreting the order code
The order code and serial number of your product can be found in the following locations:
• On the nameplate
• In the delivery papers
Obtaining information on the product
Endress+Hauser
1.
Go to the product page for your product on the Internet.
2.
At the bottom of the page, select the "Online Tools" link followed by "Check your
device features".
An additional window opens.
3.
Enter the order code from the nameplate into the search field, and then select "Show
details".
You will receive information on each feature (selected option) of the order code.
9
Incoming acceptance and product identification
4.3
Oxymax COS61D
Scope of delivery
Scope of delivery of sensor
• Oxygen sensor with transport cap or mounted cleaning system (optional)
• Operating Instructions
10
Endress+Hauser
Oxymax COS61D
Installation
5
Installation
5.1
Installation conditions
5.1.1
Orientation
45°
135°
45°
0°
180°
A0031951
2
Angle of installation
Recommended angle of inclination
The sensor must be installed at an angle of inclination in an assembly, bracket or
appropriate process connection. Recommended angle: 45°, to prevent the formation of air
bubbles. At angles of inclination of 45 to 135°, air bubbles at the oxygen-sensitive
membrane may increase the measured value.
The sensor can be installed up to the horizontal in an assembly, support or suitable process
connection. The optimum installation angle is 45°.
Other angles and upside-down installation are not recommended. Reason: Possible
sediment formation and resulting falsification of measured value.
Observe the instructions for installing sensors in the Operating Instructions for the
assembly used.
5.1.2
Mounting location
• Choose a mounting location that can be easily accessed at a later stage.
• Ensure that upright posts and fittings are fully secured and vibration-free.
• Select an installation location which produces a typical oxygen concentration for the
individual application.
Endress+Hauser
11
Installation
Oxymax COS61D
5.2
Mounting the sensor
5.2.1
Measuring system
COS61D
A complete measuring system consists of the following components at least:
• Oxymax COS61D oxygen sensor
• Liquiline CM44x multi-channel transmitter
• Sensor cable, optionally available with M12 connector
• Assembly, e.g. flow assembly COA250, immersion assembly CYA112 or
retractable assembly COA451
Optionally:
• Assembly holder Flexdip CYH112 for immersion operation
• Extension cable CYK11
• Cleaning system
1
2
6
3
5
4
A0012882
12
3
Example of a measuring system with COS61D
1
2
3
Sensor cable
Liquiline CM44x
Flexdip CYH112
4
5
6
Basin rim with railing
Oxymax COS61D
FlexdipCYA112
Endress+Hauser
Oxymax COS61D
Installation
5.2.2
Mounting the cleaning unit
If the cleaning unit is not supplied in a preassembled state:
1.
Unscrew the protection guard.
Retain the protection guard for possible reuse at a later stage without the
cleaning unit.
2.
Carefully screw on the shaft sleeve as far as it will go.
The cleaning nozzle should now be level with the spot.
1
2
1
2
Cleaning nozzle
Spot
3.
Connect the hose for the compressed air supply (to be provided onsite) or compressor
(→ 29) to the hose connection of the cleaning unit.
5.2.3
Installing a measuring point
Installation in suitable assembly is required (depending on the application)
Endress+Hauser
13
Installation
Oxymax COS61D
LWARNING
Electrical voltage
In the event of a fault, non-grounded metallic assemblies may be under voltage and then
are not touchable.
‣ When using metallic assemblies and installation equipment, observe the national
grounding provisions.
For immersion operation, install the individual modules away from the basin on a
solid base. Only carry out the final installation at the intended installation location.
Select an installation location that is easy to access.
For a complete installation of a measuring point, proceed as follows:
1.
Install a retractable or a flow assembly (if used) into the process.
2.
Connect the water supply to the rinse connections (if you use an assembly with
cleaning function).
3.
Install and connect the oxygen sensor.
NOTICE
Installation error
Cable open circuit, loss of sensor due to cable separation, unscrewing of fluorescence cap
‣ Do not install the sensor suspended from the cable.
‣ Screw the sensor into the assembly so that the cable is not twisted.
‣ When installing or uninstalling the sensor body, hold it tightly. Turn using only the
hexagonal nut on the armored coupling. Otherwise you might unscrew the
fluorescence cap. This will then remain in the assembly or process.
‣ Avoid exerting excessive tensile force on the cable (e.g. from jerky pulling).
‣ Select an installation location that is easy to access for later calibrations.
‣ Observe the instructions for installing sensors in the Operating Instructions for the
assembly used.
14
Endress+Hauser
Oxymax COS61D
Installation
5.3
Installation examples
5.3.1
Immersion operation
Universal holder and chain assembly
1
5
2
1
2
6
3
3
4
4
5
A0030564
4
1
2
3
4
5
Endress+Hauser
Chain holder on railing
Chain
Holder Flexdip CYH112
Rail
Sensor Oxymax
Wastewater assembly Flexdip CYA112
A0030565
5
1
2
3
4
5
6
Chain holder on upright post
Weather protection cover CYY101
Controller / transmitter
Chain
Wastewater assembly Flexdip CYA112
Sensor Oxymax
Holder Flexdip CYH112
15
Installation
Oxymax COS61D
Universal holder and fixed immersion tube
1
2
5
3
4
A0030567
6
1
2
3
4
5
Assembly holder with immersion tube
Protective cover
Controller / transmitter
Flexdip CYA112 immersion assembly
Sensor Oxymax
Assembly holder Flexdip CYH112
Basin rim mounting with immersion tube
1
2
3
4
A0030568
7
1
2
3
4
16
Basin rim mounting
Pendulum holder CYH112
Assembly Flexdip CYA112
Assembly float
Sensor Oxymax
Endress+Hauser
Oxymax COS61D
Installation
Float
The CYA112 float is for use in the case of large fluctuations in water level, for example in
rivers or lakes.
1
1
220 (8.66)
3
4
Fixing ring for rope and chains with terminal screw
3
Eyelets Ø15, 3 x 120 ° for anchoring
4
Plastic float, resistant to salt water
5
Pipe 40 x 1, stainless steel 1.4571
6
Bumper and ballast
7
Oxygen sensor
Ø200 (7.87)
500 (19.69)
800 (31.50)
2
Cable run with strain relief and rain shield
2
G1
5
6
7
A0032159
8
5.3.2
Dimensions in mm (inch)
Flow assembly COA250
1
3
8
7
6
5
COA250
10
1
2
3
4, 7
5
6
8
Endress+Hauser
4
A0030570
A0013319
9
2
Bypass installation with manually actuated valves or solenoid
valves
Main pipe
Medium return
Oxygen sensor
Manually actuated or solenoid valves
Flow assembly COA250-A
90 ° pipe elbow
Medium removal
17
Installation
Oxymax COS61D
5.3.3
Retractable assembly COA451
2
Detail A, 90 ° r
Detail A
3
3
1
4
5
A0030571
11
1
2
3
4
Permissible and impermissible sensor installation positions with retractable assembly COA451
Ascending pipe, best position
Horizontal pipe, sensor top down, impermissible due to air cushion or foam bubble forming
Horizontal pipe, lateral installation with permissible installation angle (acc. to sensor version)
Down pipe, impermissible
NOTICE
Sensor not fully immersed in the medium, deposit on sensor membrane or sensor
optics, sensor installed upside down
Incorrect measurements are possible and these may affect the measuring point.
‣ Do not install the assembly at points where air pockets or bubbles form or where
suspended particles may build up at the sensor membrane or sensor optics (item 2).
5.4
Post-installation check
• Are the sensor and cable undamaged?
• Is the orientation correct?
• Is the sensor installed in an assembly and is not suspended from the cable?
• Avoid the penetration of moisture by fitting the protective cap to the immersion
assembly.
18
Endress+Hauser
Oxymax COS61D
Electrical connection
6
Electrical connection
LWARNING
Device is live
Incorrect connection may result in injury or death
‣ The electrical connection may be performed only by an electrical technician.
‣ The electrical technician must have read and understood these Operating Instructions
and must follow the instructions contained therein.
‣ Prior to commencing connection work, ensure that no voltage is present on any cable.
6.1
Connecting the sensor
Connection data
• Sensor cable connected directly to the terminal connector of the basic module
• Optional: Sensor cable plug connected to the M12 sensor socket on the underside of the
transmitter
With this type of connection, the transmitter is already wired at the factory.
12
6.2
PK
GY
U+
U–
GN
YE
COM A
COM B
Sensor fixed cable with terminated cable cores
Ensuring the degree of protection
Only the mechanical and electrical connections which are described in these instructions
and which are necessary for the required, designated use, may be carried out on the device
delivered.
‣ Exercise care when carrying out the work.
Otherwise, the individual types of protection (Ingress Protection (IP), electrical safety, EMC
interference immunity) agreed for this product can no longer be guaranteed due, for
example, to covers being left off or cable (ends) which are loose or insufficiently secured.
6.3
Post-connection check
Device condition and specifications
Notes
Are the outside of the sensor, assembly, cable undamaged?
Visual inspection
Electrical connection
Notes
Are the installed cables strain-relieved and not twisted?
Is a sufficient length of the cable cores stripped, and is it
positioned in the terminal correctly?
Check the fit (by pulling gently)
Are all the screws terminals properly tightened?
Tighten
Are all cable entries mounted, tightened and leak-tight?
For lateral cable entries, make sure the
cables loop downwards to allow water to
drip off
Are all cable entries installed downwards or mounted laterally?
Endress+Hauser
19
Calibration and adjustment
Oxymax COS61D
7
Calibration and adjustment
7.1
Types of calibration
The following types of calibration are possible:
• Slope
– Air, saturated with water vapor
– Air-saturated water
– Air, variable
– Test gas calibration
– Data entry
• Zero point
– Single-point calibration in nitrogen or COY8 zero-point gel
– Data entry
• Sample calibration
Slope
• Temperature adjustment
7.2
Calibration intervals
Specifying the intervals
If you want to calibrate the sensor intermittently for a special application and/or on
account of a special type of installation, you can calculate the intervals using the following
method:
1.
Remove the sensor from the medium.
2.
Clean the outside of the sensor with a damp tissue.
3.
Then dry the sensor diaphragm carefully with a paper towel for example. (only
amperometric sensors)
4.
NOTICE
Incorrect measurements caused by atmospheric influences
‣ Protect the sensor against external influences such as sunlight and wind.
After 20 minutes, measure the oxygen saturation index in air.
5.
Decide using the results:
If the measured value is not 100 ± 2 %SAT, you must calibrate the sensor.
If the values are within the interval specified, you do not need to calibrate the
sensor. Extend the period until the next inspection.
6.
Repeat the steps specified after two, four or eight months to determine the optimum
calibration interval for your sensor.
In any case, calibrate the sensor at least once a year.
20
7.3
Calibration in air
1.
Remove the sensor from the medium.
2.
Clean the outside of the sensor with a damp tissue.
3.
Allow approx. 20 minutes for the sensor temperature to adapt to the ambient air.
Make sure that the sensor is not exposed to any direct ambient influences (direct
sunlight, drafts) during this time.
Endress+Hauser
Oxymax COS61D
Calibration and adjustment
4.
When the measured value display on the transmitter is stable, carry out the
calibration according to the operating instructions for the transmitter. Pay particular
attention to the software settings for the stability criteria for calibration and for the
ambient pressure.
5.
Where necessary:
Adjust the sensor.
6.
Then insert the sensor into the medium.
Make sure you comply with the instructions for calibration in the Operating
Instructions of the transmitter.
7.4
Calculation example for the calibration value
As a check, you can calculate the expected calibration value (transmitter display) as shown
in the following example (salinity is 0).
1.
2.
Determine:
• the sensor temperature (air temperature for the calibration type "air", water
temperature for the calibration type "air-saturated water")
• the altitude above sea level
• the current air pressure (= rel. air pressure to sea level) at the time of calibration. (If
undeterminable, use 1013 hPa (407 in H2O)).
Define:
• the saturation value S acc. to Table 1
• the altitude factor K acc. to Table 2
Table 1
T
[˚C (˚F)]
S
[mg/l=ppm]
T
[˚C (˚F)]
S
[mg/l=ppm]
T
[˚C (˚F)]
S
[mg/l=ppm]
T
[˚C (˚F)]
S
[mg/l=ppm]
0 (32)
14.64
11 (52)
10.99
21 (70)
8.90
31 (88)
7.42
1 (34)
14.23
12 (54)
10.75
22 (72)
8.73
32 (90)
7.30
2 (36)
13.83
13 (55)
10.51
23 (73)
8.57
33 (91)
7.18
3 (37)
13.45
14 (57)
10.28
24 (75)
8.41
34 (93)
7.06
4 (39)
13.09
15 (59)
10.06
25 (77)
8.25
35 (95)
6.94
5 (41)
12.75
16 (61)
9.85
26 (79)
8.11
36 (97)
6.83
6 (43)
12.42
17 (63)
9.64
27 (81)
7.96
37 (99)
6.72
7 (45)
12.11
18 (64)
9.45
28 (82)
7.82
38 (100)
6.61
8 (46)
11.81
19 (66)
9.26
29 (84)
7.69
39 (102)
6.51
9 (48)
11.53
20 (68)
9.08
30 (86)
7.55
40 (104)
6.41
10 (50)
11.25
Table 2
Altitude
[m (ft)]
Endress+Hauser
K
Height
[m (ft)]
K
Height
[m (ft)]
K
Height
[m (ft)]
K
0 (0)
1.000
550 (1800)
0.938
1050 (3450) 0.885
1550 (5090) 0.834
50 (160)
0.994
600 (1980)
0.932
1100 (3610) 0.879
1600 (5250) 0.830
100 (330)
0.988
650 (2130)
0.927
1150 (3770) 0.874
1650 (5410) 0.825
150 (490)
0.982
700 (2300)
0.922
1200 (3940) 0.869
1700 (5580) 0.820
200 (660)
0.977
750 (2460)
0.916
1250 (4100) 0.864
1750 (5740) 0.815
21
Calibration and adjustment
Oxymax COS61D
Altitude
[m (ft)]
K
Height
[m (ft)]
K
Height
[m (ft)]
K
Height
[m (ft)]
K
250 (820)
0.971
800 (2620)
0.911
1300 (4270) 0.859
1800 (5910) 0.810
300 (980)
0.966
850 (2790)
0.905
1350 (4430) 0.854
1850 (6070) 0.805
350 (1150)
0.960
900 (2950)
0.900
1400 (4600) 0.849
1900 (6230) 0.801
400 (1320)
0.954
950 (3120)
0.895
1450 (4760) 0.844
1950 (6400) 0.796
450 (1480)
0.949
1000 (3300) 0.890
1500 (4920) 0.839
2000 (6560) 0.792
500 (1650)
0.943
3.
4.
Calculate the factor L:
Relative air pressure at calibration
L=
----------------------------------------------1013 hPa
Calculate the calibration value C:
C= S.K.L
Example
• Air calibration at 18 °C (64 °F), altitude 500 m (1650 ft) above sea level, current air
pressure 1009 hPa (405 in H2O)
• S = 9.45 mg/l, K = 0.943, L = 0.996
• Calibration value C = 8.88 mg/l.
You do not need factor K in the table if your measuring device returns the absolute air
pressure L abs (air pressure depending on altitude) as the measured value. The formula
for calculation is then: C = S . Labs.
22
Endress+Hauser
Oxymax COS61D
Commissioning
8
Commissioning
8.1
Function check
Before first commissioning, check if:
• the sensor is correctly installed
• the electrical connection is correct.
If using an assembly with automatic cleaning, check that the cleaning medium (e.g. water
or air) is connected correctly.
LWARNING
Escaping process medium
Risk of injury from high pressure, high temperatures or chemical hazards
‣ Before applying compressed air to an assembly with cleaning facility, make sure the
connections are correctly fitted.
Do
not install the assembly in the process if you cannot make the correct connection
‣
reliably.
8.2
Sensor calibration
The sensor is calibrated at the factory. A new calibration is only needed in special
situations.
8.3
Cleaning the sensor automatically
Compressed air is most suitable for cyclic cleaning. The cleaning unit is either provided or
can be retrofitted, and is attached to the sensor head. It operates at a capacity of 20-60 l/
min. Optimum results are achieved at 2 bar (29 psi) and 60 l/min.
The following settings are recommended for the cleaning unit:
Endress+Hauser
Type of soiling
Cleaning interval
Cleaning duration
Media containing grease and oils
15 min
20 s
Biofilm
60 min
20 s
23
Troubleshooting
Oxymax COS61D
9
Troubleshooting
9.1
Trouble shooting instructions
If one of the following problems is present, check the measuring system in the sequence
listed here.
Problem
Testing
Remedial action
No display, no sensor reaction
Power supply to the transmitter?
‣ Connect power supply.
Sensor cable connected correctly?
‣ Establish correct connection.
Medium flow present?
‣ Create medium flow.
Deposit on the fluorescence layer? ‣
Displayed value too high
Is sensor calibrated/adjusted?
Clean sensor.
‣ Recalibrate/readjust.
Displayed temperature clearly too
low?
‣ Check sensor, if necessary send sensor
in for repair.
Has salinity value been taken into ‣
account?
Displayed value too low
Enter salinity value on transmitter.
Is sensor calibrated/adjusted?
‣ Recalibrate/readjust.
Medium flow present?
‣ Create medium flow.
Displayed temperature clearly too
high?
‣ Check sensor, if necessary send sensor
in for repair.
Deposit on the fluorescence layer? ‣
Display in Vol% or %SAT not
plausible.
During calibration, enter the
current air pressure at the
transmitter.
Medium pressure has not been
taken into account.
Clean sensor.
‣ Enter medium pressure on transmitter.
Please observe the troubleshooting information in the Operating Instructions for the
transmitter. If need be, carry out a test on the transmitter.
9.2
Check sensor
Testing
Corrective action
Set point
Slope check
‣ Move the sensor out to the air.
‣ Dry it with a paper towel.
Measured value display after 1 minute:
approx. 100 % SAT
Zero-point check ‣
Immerse the sensor in zero-point gel
(→ 28).
Display close to 0 mg/l (0 % SAT)
‣ In the case of deviations from the set points:
Follow the troubleshooting instructions or contact your sales office.
24
Endress+Hauser
Oxymax COS61D
Maintenance
10
Maintenance
Take all the necessary precautions in time to ensure the operational safety and reliability
of the entire measuring system.
NOTICE
Effects on process and process control
‣ When carrying out any work on the system, take into account possible repercussions
for process control or the process itself.
‣ For your own safety, only use genuine accessories. With genuine parts, the function,
accuracy and reliability are also ensured after maintenance work.
10.1
Maintenance schedule
Maintenance cycles depend to a great extent on the operating conditions.
The following rule of thumb applies:
• Constant conditions, e.g. power plant = long cycles (1/2 year)
• Greatly varying conditions, e.g. daily CIP cleaning = short cycles (1 month or shorter)
The following method helps you determine the necessary intervals:
1.
Inspect the sensor one month after commissioning. Take it out of the medium and
dry it.
2.
After 10 minutes, measure the oxygen saturation index in air.
Decide depending on the result: If the measured value is not 100 ±2 %SAT, you
must maintain the sensor. Otherwise, double the length of time to the next
inspection.
3.
Proceed as per Point 1 after two, four and/or eight months. In this way, you can
determine the optimum maintenance interval for your sensor.
Particularly in the case of widely fluctuating process conditions, damage may occur to
the fluorescence layer even within a maintenance cycle. You can recognize this by
implausible sensor behavior.
10.2
Maintenance tasks
The following tasks are mandatory:
• Clean the sensor and of fluorescence cap
• Replacement of wear parts or consumables:
– Fluorescence cap
– Sealing ring
• Check the measuring function:
1. Remove the sensor from the medium.
2. Clean and dry the fluorescence cap.
3. After about 10 minutes, measure the oxygen saturation index in air (without
recalibration).
4. The measured value should be 100 ± 2 % SAT.
• Recalibration (if desired or required)
10.3
Clean exterior of sensor.
The measurement can be corrupted by sensor fouling or malfunction due to the following,
for example:
Deposit on the fluorescence cap
These result in longer response times and, under certain circumstances, a reduced
slope.
Endress+Hauser
25
Maintenance
Oxymax COS61D
For reliable measurement, the sensor must be cleaned at regular intervals. The frequency
and intensity of the cleaning operation depend on the measuring medium.
Clean the sensor:
• before every calibration
• at regular intervals during operation as necessary
• before returning it for repairs.
Type of soiling
Cleaning
Salt deposits
Immerse the sensor in drinking water or in 1-5% hydrochloric acid
(for a few minutes). Afterwards, rinse it with copious amounts of
water.
Dirt particles on the sensor shaft and
shaft sleeve (not fluorescence cap!)
Clean the sensor shaft and sleeve with water and use a suitable
sponge.
Dirt particles on fluoresence cap
Clean the fluorescence cap with water and a soft tissue.
After cleaning, rinse the sensor with copious amounts of clean water.
10.4
Cleaning the sensor optics
The optics need to be cleaned only if medium has penetrated through a defective
fluorescence cap.
1.
Unscrew the protection guard and fluorescence cap from the sensor head.
2.
Carefully clean the optical surface with a soft tissue until the deposit is fully removed.
3.
Clean the optics with drinking water or distilled water.
4.
Dry the optics and screw on a new fluorescence cap.
NOTICE
Damage, scratches on optical surface
Distorted measured values
‣ Make sure that the optical surface is not scratched or damaged in any other way.
10.5
Wear parts and consumables
Parts of the sensor are subject to wear during operation. By taking suitable measures, you
can restore the normal operating function.
Corrective action
Reason
Process seals
visible damage to a process seal
Replace fluorescencecap
Fluorescence layer can no longer be cleaned or is damaged (hole or overstretching)
10.5.1
Replacing sealing rings
The sealing ring must be replaced if visibly damaged. For replacement, use only original
sealing rings.
10.5.2
Replacing fluorescence cap
The typical operating life of a fluorescence cap is more than 2 years. The sensor checks
whether the cap is aging and issues a warning via the transmitter if the rate of aging
reaches a specific value. The sensor is still able to measure at this stage. However, it is
advisable to change the cap as quickly as possible.
26
Endress+Hauser
Oxymax COS61D
Maintenance
Removing the old fluorescence cap
1.
Remove the sensor from the medium.
2.
Unscrew the protection guard.
3.
Clean the outside of the sensor.
4.
Unscrew the fluorescence cap.
5.
Clean and dry the optical surface.
Installing the new fluorescence cap
Make sure that there are no dirt particles on the sealing surfaces.
Endress+Hauser
6.
Carefully screw the fluorescence cap onto the sensor head until the stop.
7.
Screw the protection guard back on.
After replacing the fluorescence cap, you must recalibrate and adjust the sensor.
8.
Then insert the sensor into the medium and check that no alarm is displayed on the
transmitter.
27
Accessories
Oxymax COS61D
11
Accessories
The following are the most important accessories available at the time this
documentation was issued. For accessories not listed here, please contact your service
or sales office.
11.1
Assemblies (selection)
Flexdip CYA112
• Immersion assembly for water and wastewater
• Modular assembly system for sensors in open basins, channels and tanks
• Product Configurator on the product page: www.endress.com/cya112
Technical Information TI00432C
Flowfit COA250
• Flow assembly for oxygen measurement
• Product Configurator on the product page: www.endress.com/coa250
Technical Information TI00111C
Cleanfit COA451
• Manual retractable assembly made of stainless steel with ball valve shutoff
• For oxygen sensors
• Product Configurator on the product page: www.endress.com/coa451
Technical Information TI00368C
11.2
Assembly holder
Flexdip CYH112
• Modular holder system for sensors and assemblies in open basins, channels and tanks
• For Flexdip CYA112 water and wastewater assemblies
• Can be affixed anywhere: on the ground, on the capstone, on the wall or directly onto
railings.
• Stainless steel version
• Product Configurator on the product page: www.endress.com/cyh112
Technical Information TI00430C
11.3
Measuring cable
Memosens data cable CYK11
• Extension cable for digital sensors with Memosens protocol
• Product Configurator on the product page: www.endress.com/cyk11
Technical Information TI00118C
11.4
Zero-point gel
COY8
Zero-point gel for oxygen sensors
• Oxygen-free gel for the validation, calibration and adjustment of oxygen measuring cells
• Product Configurator on the product page: www.endress.com/coy8
Technical Information TI01244C
28
Endress+Hauser
Oxymax COS61D
Accessories
11.5
Protection guard
Membrane basket protector COY3-SK
• For using the sensor in fish farming tanks
• Order No.: 50081787
11.6
Cleaning unit
Compressed air cleaning for COSXX
• Connection: AD 6/8 mm (incl. hose reducer coupling) or AD 6.35 mm (¼")
• Materials: POM/V4A
• Order No.
– AD 6/8 mm: 71110801
– AD 6.35 mm (¼"): 71110802
Compressor
• For compressed air cleaning
• Order No.
– 230 V AC order no. 71072583
– 115 V AC order no. 71194623
11.7
Calibration vessel
Calibration vessel
• For COS61D/61
• Order No.: 51518599
Endress+Hauser
29
Repair
Oxymax COS61D
12
Repair
12.1
Spare parts and consumables
1
Item
Spare parts kit
Order number
1
Sensor
As per product structure
2
Sealing ring x 2
51518597
3
Sensor cap (fluorescence cap)
51518598
4
Protection guard
50053276
2
3
4
12.2
Return
The product must be returned if repairs or a factory calibration are required, or if the
wrong product was ordered or delivered. As an ISO-certified company and also due to legal
regulations, Endress+Hauser is obliged to follow certain procedures when handling any
returned products that have been in contact with medium.
To ensure swift, safe and professional device returns, please read the return procedures
and conditions at www.endress.com/support/return-material.
12.3
Disposal
The device contains electronic components and must therefore be disposed of in
accordance with regulations on the disposal of electronic waste.
Observe the local regulations.
30
Endress+Hauser
Oxymax COS61D
Technical data
13
Technical data
13.1
Input
Measured values
Dissolved oxygen [mg/l, µg/l, ppm, ppb or % SAT or hPa]
Measuring ranges
Measuring ranges apply for 20 ˚ (68 ˚F) and 1013 hPa (15 psi)
with Liquiline CM44x, CM44xR:
• 0 to 20 mg/l
• 0 to 200 % SAT
• 0 to 400 hPa
13.2
Performance characteristics
Response time
From air to nitrogen at reference operating conditions:
t90 : 60 s
reference operating
conditions
Reference temperature:
Reference pressure:
Reference application:
25 ˚C (77 ˚F)
1013 hPa (15 psi)
Air-saturated water
Maximum measured error
Measuring range
< 12 mg/l
12 mg/l to 20 mg/l
Maximum measured error
0.01 mg/l or ±1 % of reading
±2% of reading
Repeatability
±0.5 % of end of measuring range
Operating life of sensor cap
>2 years (under reference operating conditions, protect against direct sunlight)
13.3
Environment
Ambient temperature
range
-20 to +60 °C (-4 to 140 °F)
Storage temperature
-20 to +70 °C (0 to 160 °F)
at 95% relative air humidity, not condensating
Degree of protection
• Fixed cable with end sleeves:
IP 68 (test conditions: 10 m (33 ft) water column at 20 °C (68 °F) over 7 days)
• Fixed cable with M12 connector:
IP 68 (test conditions: 1 m (3.3 ft) water column, 3N KCl, at 50 °C (122 °F) over 30 days)
Electromagnetic
compatibility
Interference emission and interference immunity as per EN 61326: 2005, Namur NE
21:2007
Endress+Hauser
31
Technical data
Oxymax COS61D
13.4
Process
Process temperature
–5 to +60 ˚C (20 to 140 ˚F)
Process pressure
Ambient pressure ... 10 bar (... 145 psi) abs.
13.5
Mechanical construction
21 (0.83)
15 (0.59)
Ø40 (1.57)
A0013311
13
With optional M12 connector
Dimensions in mm (inch)
32
242 (9.53)
196 (7.72)
11 (0.43)
NPT 3/4"
G1
77 (3.04) / 66 (2,59)*
230 (9.06)
186 (7.32)
12 (0.47)
47 (1.85)
97 (3.82)
Dimensions
A0013312
14
*
With optional cleaning unit
depending on version of cleaning unit
Endress+Hauser
Oxymax COS61D
Technical data
50 (1.97)
77 (3.04)
11 (0.43)
19 (0.75)
Optional cleaning unit
.57)
Ø40 (1
19 (0.75)
24°
A
28.5 (1.12)
B
61 (2.40)
A0013314
15
Dimensions in mm (inch)
Weight
with cable length 7 m (23 ft):
with cable length 15 m (49 ft):
with TOP68 plug connection:
Materials
Parts in contact with medium
Sensor shaft
Cap with fluorescence layer
Fluorescence layer
0.7 kg (1.5 lbs)
1.1 kg (2.4 lbs)
0.3 kg (0.66 lbs) depending on version
Process connection
G1, NPT 3/4"
Sensor cable
Shielded 4-core fixed cable
Cable connection at
transmitter
• Terminal connection, end sleeves
• Optional: M12 connector
Maximum cable length
max. 100 m (330 ft), incl. Cable extension
Temperature compensation
Internal
Interface
Memosens protocol
Endress+Hauser
Stainless steel 1.4435 (AISI 316L)
POM
Silicone
33
Index
Index
A
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Ambient temperature range . . . . . . . . . . . . . . . . . . .
Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
28
20
31
28
C
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . 31
Ensuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Device description . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
E
Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . 19
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
F
Fluorescence cap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Replace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
I
Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . . 9
Installation
Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Cleaning unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Installation instructions . . . . . . . . . . . . . . . . . . . . . . . 11
34
O
Occupational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Operating life of sensor cap . . . . . . . . . . . . . . . . . . . . 31
Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Optical measuring principle . . . . . . . . . . . . . . . . . . . . . 6
Orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
P
Performance characteristics . . . . . . . . . . . . . . . . . . . . 31
Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Process connection . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Process pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Process temperature . . . . . . . . . . . . . . . . . . . . . . . . . 32
Product identification . . . . . . . . . . . . . . . . . . . . . . . . . 9
Product safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
R
D
Maintenance schedule . . . . . . . . . . . . . . . . . . . . . . . .
Maintenance tasks . . . . . . . . . . . . . . . . . . . . . . . . . .
Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Maximum measured error . . . . . . . . . . . . . . . . . . . . .
Measured values . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measuring cable . . . . . . . . . . . . . . . . . . . . . . . . . . . .
N
Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Calibration
Calculation example . . . . . . . . . . . . . . . . . . . . . . . 21
In air . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Types of calibration . . . . . . . . . . . . . . . . . . . . . . . 20
Check
Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Cleaning
Sensor optics . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Cleaning the sensor optics . . . . . . . . . . . . . . . . . . . . . 26
Cleaning unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Connection
Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Ensuring the degree of protection . . . . . . . . . . . . . 19
Cust. tag number . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
M
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Measuring ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Mode of operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
25
25
33
31
31
28
reference operating conditions . . . . . . . . . . . . . . . . . .
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Replacing sealing rings . . . . . . . . . . . . . . . . . . . . . . .
Response time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
31
30
31
26
31
30
S
Safety
Occupational safety . . . . . . . . . . . . . . . . . . . . . . . . 4
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Product . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Sensor
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23, 25
Connecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Sensor design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . 31
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
T
Technical data
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Mechanical construction . . . . . . . . . . . . . . . . . . . .
Performance characteristics . . . . . . . . . . . . . . . . .
31
31
32
31
Endress+Hauser
Index
Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Trouble shooting instructions . . . . . . . . . . . . . . . . . . 24
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
U
Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
W
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Wear parts and consumables . . . . . . . . . . . . . . . . . . . 26
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Z
Zero-point gel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Endress+Hauser
35
*71341281*
71341281
www.addresses.endress.com