Endress + Hauser Operating Instructions TOCII CA72TOC Analyzer for online determination of TOC
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
BA00448C/07/EN/13.15
71305431
Products
Solutions
Services
Operating Instructions
TOCII CA72TOC
Analyzer for online determination of TOC in aqueous
media using thermic catalytic combustion
Document information
Warnings
The structure, signal words and color coding applied to safety information comply with the
requirements of ANSI Z535.6 ("Product safety information in product manuals, instructions
and other collateral materials").
Structure of information
!
DANGER
Cause (/consequences)
Consequences of noncompliance (if applicable)
‣ Corrective action
!
WARNING
Cause (/consequences)
Consequences of noncompliance (if applicable)
‣ Corrective action
!
CAUTION
Cause (/consequences)
Consequences of noncompliance (if applicable)
‣ Corrective action
NOTICE
Cause/situation
Consequences of noncompliance (if applicable)
‣ Action/note
Meaning
This symbol alerts you to a dangerous situation.
If you do not avoid this dangerous situation, it will result in
death or serious injury.
This symbol alerts you to a dangerous situation.
If you do not avoid the dangerous situation, it may result in
death or serious injury.
This symbol alerts you to a dangerous situation.
Failure to avoid this situation can result in minor or more
serious injuries.
This symbol alerts you to situations which may result in
damage to property.
Symbols used
Additional information, tips
Permitted or recommended
Forbidden or not recommended
Endress+Hauser
TOCII CA72TOC
Table of contents
1
Basic safety instructions . . . . . . . . . . . . . 5
6
Commissioning . . . . . . . . . . . . . . . . . . . 34
1.1
1.2
1.3
1.4
1.5
Personnel requirements . . . . . . . . . . . . . . . . . . . . . . 5
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Workplace safety . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Product safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
6.1
6.2
2
Incoming acceptance and product
identification . . . . . . . . . . . . . . . . . . . . . . 7
2.1
2.2
2.3
2.4
Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . 7
Product identification . . . . . . . . . . . . . . . . . . . . . . . . 7
2.2.1 Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.2.2 Product identification . . . . . . . . . . . . . . . . . 7
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Certificates and approvals . . . . . . . . . . . . . . . . . . . . 8
3
Installation . . . . . . . . . . . . . . . . . . . . . . . . 9
3.1
3.3
Installation conditions . . . . . . . . . . . . . . . . . . . . . . . 9
3.1.1 Methods of installation . . . . . . . . . . . . . . . . 9
3.1.2 Compressed air and water supply . . . . . 10
3.1.3 Gas flow . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.1.4 Environment . . . . . . . . . . . . . . . . . . . . . . . 10
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2.1 Mounting sequence . . . . . . . . . . . . . . . . . 11
3.2.2 Wall mounting with a pivoting frame . . 11
3.2.3 Mounting on a base frame . . . . . . . . . . . 12
3.2.4 Mounting the CO2 adsorber . . . . . . . . . . 12
3.2.5 Connecting the media . . . . . . . . . . . . . . . 13
3.2.6 Preparing the chemicals . . . . . . . . . . . . . 15
Post-installation check . . . . . . . . . . . . . . . . . . . . . 19
6.5
Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
6.2.1 Commissioning sequence . . . . . . . . . . . . . 34
6.2.2 Updating the analyzer software . . . . . . . . 35
Programming mode . . . . . . . . . . . . . . . . . . . . . . . . 37
6.3.1 Menu SETTING - RANGE DATA . . . . . . . 37
6.3.2 Menu SETTING - BASIC DATA . . . . . . . . . 39
6.3.3 Menu SETTING - ALARM LIMITS . . . . . . 40
6.3.4 Menu SETTING - SET CLOCK . . . . . . . . . . 41
6.3.5 Menu SETTING - SET BRIGHTN./
CONTR. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.3.6 Menu SETTING - MEASURING SITE . . . . 42
6.3.7 Menu LISTS - SHOW COMPLETE
RECORDS . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.3.8 Menu LISTS - ALARM RECORD . . . . . . . . 43
6.3.9 Menu LISTS - MAINTENANCE RECORD . 45
6.3.10 Menu LISTS - MIN MAX AVERAGE . . . . 47
6.3.11 Menu LISTS - RECORD ALL DATA . . . . . . 47
6.3.12 Menu TEST - INPUT TEST . . . . . . . . . . . . 49
6.3.13 Menu TEST - OUTPUT TEST . . . . . . . . . . . 50
6.3.14 Two-channel: external changeover . . . . . 51
6.3.15 Two-channel: time-controlled
changeover . . . . . . . . . . . . . . . . . . . . . . . . . 53
Optimizing the analyzer . . . . . . . . . . . . . . . . . . . . . 55
6.4.1 Adjusting the measuring range . . . . . . . . 55
6.4.2 Influence of dosing volume (pump P2)
on the measuring signal . . . . . . . . . . . . . . 55
6.4.3 Operating the analyzer with high salt
loads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
4
Electrical connection . . . . . . . . . . . . . . . 20
7
Diagnostics and troubleshooting . . . . 60
4.1
7.1
Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
4.2
Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.1.1 Preparing to wire . . . . . . . . . . . . . . . . . . .
4.1.2 Power distribution . . . . . . . . . . . . . . . . . .
4.1.3 Signal connection . . . . . . . . . . . . . . . . . . .
4.1.4 Power unit . . . . . . . . . . . . . . . . . . . . . . . .
4.1.5 Distributor connection . . . . . . . . . . . . . .
4.1.6 Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Post-connection check . . . . . . . . . . . . . . . . . . . . .
5
Operation. . . . . . . . . . . . . . . . . . . . . . . . . 28
5.1
5.2
5.3
Operating elements . . . . . . . . . . . . . . . . . . . . . . .
Display and keypad . . . . . . . . . . . . . . . . . . . . . . . .
Operation during measuring mode . . . . . . . . . . .
5.3.1 Recording mode . . . . . . . . . . . . . . . . . . . .
5.3.2 Spot view . . . . . . . . . . . . . . . . . . . . . . . . . .
Standby mode . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.2
5.4
Endress+Hauser
20
20
21
23
25
26
26
27
6.3
6.4
28
29
31
31
32
33
3
TOCII CA72TOC
8
Maintenance . . . . . . . . . . . . . . . . . . . . . .65
8.1
8.2
8.3
8.4
8.5
Cleaning the housing . . . . . . . . . . . . . . . . . . . . . . . 69
Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Topping up reagents . . . . . . . . . . . . . . . . . . . . . . . 70
Replacing the standard . . . . . . . . . . . . . . . . . . . . . 71
Entering the concentration of the standard
solution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
8.6 Service menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
8.7 Pumps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
8.7.1 Replacing the hoses on pumps P1
and P4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
8.7.2 Replacing the hose on pump P2 . . . . . . . 83
8.7.3 Replacing the hose on pump P3 . . . . . . . 87
8.7.4 Adjusting pump P2 . . . . . . . . . . . . . . . . . . 92
8.7.5 Cleaning dilution water pump P5 . . . . . . 96
8.8 Adjusting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
8.8.1 Adjusting the analyzer . . . . . . . . . . . . . . . 97
8.8.2 Calibrating the analyzer . . . . . . . . . . . . . 100
8.8.3 Empty volume dosing . . . . . . . . . . . . . . . 102
8.8.4 Adjustment pH sensor . . . . . . . . . . . . . . 105
8.9 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
8.9.1 Screen flush . . . . . . . . . . . . . . . . . . . . . . . 109
8.9.2 Power flush . . . . . . . . . . . . . . . . . . . . . . . 111
8.9.3 Bypass screen . . . . . . . . . . . . . . . . . . . . . . 113
8.9.4 Strip and separation chamber . . . . . . . . 115
8.9.5 Opening the circuit . . . . . . . . . . . . . . . . . 119
8.9.6 Combustion pipe . . . . . . . . . . . . . . . . . . . 121
8.9.7 Leakage test . . . . . . . . . . . . . . . . . . . . . . . 128
8.10 Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131
8.10.1 Replacing the acid filter . . . . . . . . . . . . . 131
8.10.2 Replacing the gas filter . . . . . . . . . . . . . . 133
8.10.3 Replacing the O2 filter . . . . . . . . . . . . . . 135
8.10.4 Replacing the heated salt trap . . . . . . . . 137
8.10.5 Replacing the filter pads in the
ventilators . . . . . . . . . . . . . . . . . . . . . . . . 141
9
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . .142
9.1
9.2
9.3
9.4
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
Decommissioning . . . . . . . . . . . . . . . . . . . . . . . . . 145
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
10
Accessories . . . . . . . . . . . . . . . . . . . . . .147
11
Technical Data . . . . . . . . . . . . . . . . . . .148
11.1
11.2
11.3
11.4
11.5
11.6
11.7
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148
Performance characteristics . . . . . . . . . . . . . . . . 149
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149
Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149
Mechanical construction . . . . . . . . . . . . . . . . . . . 149
Index. . . . . . . . . . . . . . . . . . . . . . . . . . . .150
4
Endress+Hauser
TOCII CA72TOC
Basic safety instructions
1
Basic safety instructions
1.1
Personnel requirements
‣ Installation, commissioning, operation and maintenance of the measuring system may
‣
‣
‣
‣
be carried out only by specially trained technical personnel.
Trained personnel must be authorized for the specified activities by the system operator.
Electrical connection must only be carried out by a certified electrician.
Technical personnel must have read and understood these Operating Instructions and
must adhere to them.
Measuring point faults may only be rectified by authorized and specially trained
personnel.
Repairs not described in the Operating Instructions provided may only be carried out
directly by the manufacturer or by the service organization.
1.2
Designated use
The analyzer is a compact thermo-catalytic analytical measurement system. It is designed to
monitor the TOC content of industrial and municipal wastewater.
The device is particularly suitable for the following applications:
• Monitoring of industrial wastewater, in the inlet and outlet
• Control of process wastewater
• Monitoring of surface run-off in industrial systems
• Monitoring of surface run-off in airports
• Municipal wastewater monitoring
• Recording the carbon load for nutrient dosing
Any other use than the one described here compromises the safety of persons and the entire
measuring system and is not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.
NOTICE
Applications which do not comply with specifications
Possibility of incorrect measurements and faults, including failure of the measuring point
‣ Use the product only in accordance with the its specification.
‣ Pay attention to the technical data on the nameplate.
1.3
Workplace safety
As the user, you are responsible for complying with the following safety conditions:
• Installation instructions
• Local prevailing standards and regulations.
1.4
Operational safety
‣ Before commissioning the entire measuring point, check all the connections. Ensure that
‣
‣
‣
Endress+Hauser
electrical cables and hose connections are not damaged.
Do not operate damaged products and secure them against unintentional
commissioning. Mark the damaged product as being defective.
If faults can not be rectified, the products must be taken out of service and secured
against unintentional commissioning.
Do not touch any rotating pump parts.
5
Basic safety instructions
TOCII CA72TOC
1.5
Product safety
The product is designed to meet state-of-the-art safety requirements, has been tested and
left the factory in a condition in which it is safe to operate.
The relevant regulations and standards have been met.
6
Endress+Hauser
TOCII CA72TOC
Incoming acceptance and product identification
2
Incoming acceptance and product
identification
2.1
Incoming acceptance
• Make sure the packaging is undamaged!
• Inform the supplier about any damage to the packaging. Keep the damaged packaging
until the matter has been settled.
• Make sure the contents are undamaged!
• Inform the supplier about damage to the contents. Keep the damaged products until the
matter has been settled.
• Check the delivery for completeness. Check it against the delivery papers and your order.
• The packaging material used to store or to transport the product must provide shock
protection and humidity protection. The original packaging offers the best protection.
Also, keep to the approved ambient conditions (see "Technical data").
• If you have any questions, please contact your supplier or your local sales center.
2.2
Product identification
2.2.1
Nameplate
The nameplate bears the following information:
• Manufacturer data
• Order code (device version)
• Serial number
• Measuring range
• Outputs and communication
• Power supply
• Degree of protection
• (Permitted) ambient conditions
Compare the data on the nameplate with your order.
2.2.2
Product identification
To establish your product version, enter the order code from the nameplate into the
search field at the following address:
www.products.endress.com/order-ident
Endress+Hauser
7
Incoming acceptance and product identification
2.3
TOCII CA72TOC
Scope of delivery
The scope of delivery comprises:
• 1 analyzer CA72TOC
• Accessories for leak check
• Tool kit to remove the glass ball and the media
• Accessories for acid filter
• Kit for commisioning of stripper and separation chamber
• Accessories for combustion furnace maintenance
• Hose kit
• 1 canister 5 liter
• 2 canisters 2 liter
• Key for door lock
• Connection box EMV
• Measuring glas 10 ml
• Sponge
• Safety glasses
• Gloves, acid-proof and base proof
• Protective gloves, heat-resistant
• Silicone grease
• Operating Instructions
If you have any questions, please contact your supplier or your local sales center.
2.4
Certificates and approvals
Declaration of conformity
The product meets the requirements of the harmonized European standards.
It thus complies with the legal requirements of the EC directives.
The manufacturer confirms successful testing of the product by affixing the 4 symbol.
CSA C/US General Purpose (optional)
The device meets the requirements of "Class 8721 06, laboratory equipment, electrical; Class
8721 86, electrical equipment for laboratory use - certified to US standards" for indoor use.
Certificate no.: 2577401
Electrical safety
In compliance with IEC 61010-1, protection class I, Installation category II.
Fluctuations in the supply voltage may not exceed 10 percent of the nominal voltage.
8
Endress+Hauser
TOCII CA72TOC
Installation
3
Installation
3.1
Installation conditions
3.1.1
Methods of installation
The analyzer can be mounted in three ways:
• As a standalone desktop unit
• Mounted to the wall (option F1)
• Mounted on a base frame (option F2)
a0023087
Fig. 1:
Dimensions in mm (inch)
Mount the device in such a way that it can also be accessed from the rear for
maintenance purposes.
Endress+Hauser
9
Installation
TOCII CA72TOC
3.1.2
Compressed air and water supply
When operating the analyzer, use only CO2-free air.
The air must be dry and oil-free and must meet the following conditions:
• < 3 ppm CO2
• < 3 ppm hydrocarbons
• The pressure must be 2 bar (29 psi) constantly.
• The pressure tolerance is ±5 %.
• The required quantity of compressed air is:
– 600 l/h (21.2 ft3/hr) for the CO2 adsorber gas generator (Domnick Hunter)
– 60 l/h (2.12 ft3/hr) for the soda-lime CO2 scrubbers
• The compressed air supply must be fitted with a CO2 scrubber (inlet pressure 4 to 10 bar
(58 to 145 psi) and a pressure regulator.
• Connection: 4/6 mm DN
The water is connected via a 6/8 mm DN or G3/8 coupling. The pressure must be between
2 and 4 bar (29 to 58 psi) (for version with "sample predilution": 3 ±0.2 bar (43.5 ±3 psi)).
In the version with sample predilution, use de-ionized water (DI water) or drinking
water with a water hardness level < 10 °dH (< 179 ppm CaCO3) for the dilution water.
3.1.3
Gas flow
When operating the analyzer, use only CO2-free air. Primary pressure at the
manometer: 2 bar (29 psi).
Circuit gas
The flowmeter for the circuit gas is used to perform function checks and is set at the factory.
The flow rate during operation is 0.7 to 1.2 l/min (1.5 to 2.5 ft3/hr).
Carrier gas
The volume flow for the carrier gas is regulated using a precision restrictor. The flow rate is
approx. 0.8 l/min (1.7 ft3/hr) at a primary pressure of 2 bar (29 psi).
Stripping gas
The volume flow for the stripping gas is also regulated using a precision restrictor. The flow
rate is approx. 0.15 l/min (0.3 ft3/hr) at a primary pressure of 2 bar (29 psi).
3.1.4
Environment
• The analyzer requires a drain beneath the device. Use a 6/8 mm drain pipe made of PTFE.
No backpressure should form in the drain.
• An exhaust gas connection is required in enclosed spaces. Halogens or other vapors may
not build up in this area. No backpressure should form in the 4/6 mm exhaust gas tube.
• Ambient temperature range: 5 to 35 °C (41 to 95 °F)
• Humidity: 20 to 80 %, non-condensating
• Pollution degree 2
• The analyzer must be accessible from the front and rear.
• 2000 m (6500 ft) above sea level
• No direct sunlight
10
Endress+Hauser
TOCII CA72TOC
Installation
3.2
!
Installation
WARNING
Danger! Risk of electric shock!
‣ Do not connect the analyzer to the power grid before all the installation work has been
completed and the liquid and gas media are connected.
‣ To connect the analyzer to the electricity supply, follow the instructions in the "Electrical
connection" section .
3.2.1
Mounting sequence
1.
Mount the analyzer on the base frame, on a table or in the pivoting frame.
2.
Mount the EMC junction box (right-hand side of device).
3.
Mount the reagent tray under the analyzer.
4.
Mount the vent on the sample conditioning system (PA-2 / PA-3 or PA-9 only).
5.
Connect the media (see the "Connecting the media" section).
6.
Establish the electrical connections (see the "Wiring" section).
3.2.2
Wall mounting with a pivoting frame
In the "Wall-mount" version, the analyzer is mounted to the wall with a pivoting frame.
a0014227
Fig. 2:
Pivoting frame for wall-mounting, dimensions in mm (inch)
All bore holes for wall-mounting have a diameter of 8.5 mm (0.33").
Endress+Hauser
11
Installation
TOCII CA72TOC
3.2.3
Mounting on a base frame
In the "Base frame" version, the analyzer is mounted on a base frame.
Mount the device in such a way that it is also accessible from the rear for maintenance
purposes.
.4)
62
0(
24
.4)
812 (32.0)
62
24
0(
mm (inch)
a0013913
Fig. 3:
3.2.4
Mounting on a base frame
Mounting the CO2 adsorber
• Position the CO2 adsorber gas generator (Domnick Hunter) on the ground or mount it on
the wall in accordance with the enclosed drawing.
• Mount the soda-lime CO2 scrubber in accordance with the enclosed Operating Instructions
BA01243C/07/EN
12
Endress+Hauser
TOCII CA72TOC
Installation
3.2.5
Connecting the media
1
12
2
3
4
5
6
11
10
9
8
7
a0012643
Fig. 4:
1
2
3
4
5
6
7
8
9
10
11
12
Endress+Hauser
Connections on the left side panel
Power cable
Gas outlet
Gas connection
Water connection
Bypass rinse water
External grounding
Standard C2 connection
Standard C1 connection
Acid connection
Condensate drainage
Sample supply
Sample drainage
13
Installation
TOCII CA72TOC
Sample supply
Sample is supplied via a 3/8" connection on the left side panel (item 11).
Sample conditioning
Inlet connection (mm / inch)
Outer diameter
Drain connection (mm / inch)
Outer diameter
PA2
40 / 1.57
50 / 1.97
PA3
20 / 0.79
30 / 1.18
PA9
20 / 0.79
32 / 1.26
Analyzer sample drainage
Sample is drained off unpressurized via a DN 6/8 mm hose connection (compression fitting)
on the left side panel (item 12) into an open channel or pipe. Route the hose in such a way
that backpressure cannot form.
Condensate drainage
Condensate is drained off without pressure via a hose gland (PE,
DN 1.6/3.2 mm, scope of delivery) on the left side panel (item 10):
• into a collecting vessel,
• into an open channel or
• into a pipe.
The condensate discharge is acidic (pH value = 2 to 2.5).
Route the hose in such a way that backpressure cannot form.
Acid connection
Place the acid cistern in the reagent tray. Connect the acid hose to the left side panel
(item 9).
Connecting the standards
Put the standard containers in the holders on the left side panel.
Connect the standards to the left side panel (C1 to item 8 and C2 to item 7).
Gas outlet
Gas escapes via a hose gland (DN 4/6 mm) on the left side panel (item 2). Ensure that there
is adequate ventilation in the room, or remove the exhaust gas from the room via a hose
(DN 4/6 mm). The end of the hose must be positioned in such a way that it is not under
pressure and is protected from frost.
14
Endress+Hauser
TOCII CA72TOC
Installation
3.2.6
!
Preparing the chemicals
WARNING
Accidents with chemicals can kill, blind, burn, or damage lungs.
Many chemicals are poisonous or corrosive; some are explosive, by themselves or in
combination. Others are hazardous because they can be absorbed easily through skin or the
respiratory system.
‣ When working with any chemicals, always observe the precautions in the section
"Precautions regarding use of chemicals".
The analyzer requires the following chemical solutions:
• Stripper solution
Nitric acid HNO3 (CAS: 7697-37-2) conc. 25 %
Once it has been diluted accordingly, this acid is used to acidify the sample. This converts
the carbonate ion CO32 into CO2 and removes the dissolved CO2 from the solution (TIC
stripping).
• Parent solution
Potassium hydrogen phthalate (KHP (CAS: 877-24-7)) with a concentration of
5 000 mg TOC/l
This parent solution for calibrating and adjusting the analyzer is used as a diluted standard
in the measuring range from 0 to 600 mg TOC/l. In the event of high KHP concentrations
and sample acidification (pH < 2.5), there is the risk of KHP precipitating in the solution.
• Parent solution
Citric acid (CAS: 5949-29-1) with a concentration of 100 000 mg TOC/l
This parent solution for calibrating and adjusting the analyzer is used as a diluted standard
in the measuring range from 600 mg TOC/l onwards.
The stripper solution and the parent solutions are available from Endress+Hauser (see
"Accessories" chapter).
Specifying the standard concentrations
The right choice of standard concentration is key to the accuracy of the measurement
method.
Before specifying the concentrations of the standard solutions, determine the concentration
range in which the device is expected to measure. The most common concentrations should
be covered by the standard solutions.
Please note, however, that there should be a concentration ratio of between 1:4 and 1:20
between the two standard solutions.
If a limit value is to be monitored, it makes sense for the concentration of a standard to
correspond to the concentration of the limit value, as this guarantees the greatest precision
when monitoring.
Please note that there is no method of TOC measurement that can provide accurate
measurement results over a span greater than 1:30.
Example:
Concentration to be measured:
Most frequent concentration:
Limit value to be monitored:
3 to 300 mg/l
50 to 150 mg/l
200 mg/l
20 and 200 mg/l should be selected here as the standard solutions. The system can then
measure accurately in the range from 10 to 300 mg/l (taking the measuring range of the
system into account). A higher measured value deviation can be expected below a
concentration level of 10 mg/l and above a concentration level of 300 mg/l.
Endress+Hauser
15
Installation
TOCII CA72TOC
Precautionary measures when working with chemicals
• Carefully read the "Material Safety Data Sheet" (MSDS) that is supplied with every chemical
to determine the hazards posed and what precautionary measures have to be taken. If you
are not sure what risks are associated with the chemical, please ask appropriately qualified
experts.
• Never prepare chemicals on your own. You may need assistance in the event of an
accident. For this reason, always make sure that someone is close-by. Only prepare
chemicals in a properly equipped laboratory.
• When preparing chemicals always wear protective goggles, rubber gloves and a rubber
apron. In addition, wear a dust mask or face shield when working with fine-powder
chemicals.
• Chemicals or solutions that are used with the analyzer (or another process analysis device)
should never be inhaled, tasted or swallowed.
• When preparing solutions, you must always label the containers appropriately, indicating
the contents and the date of preparation. Dispose of unlabeled or expired solutions in
accordance with local regulations and guidelines.
• Some chemicals are very reactive when combined with other substances or dissolved in
water and cause explosions, fire or heat to occur, or dangerous gases, foams or sprays to
be released. Avoid the risk of dangerous accidents occurring: If you do not know how
certain chemicals react, make sure you do not mix them with other substances. Never mix
chemicals that are known to react severely.
Reagent quality
Reagent quality can affect the accuracy of measurements. Only use high-quality
reagents. For this reason, follow the steps outlined below to ensure reagent quality.
• For 10, 100, and 500 mg/l units, use the best grade of reagents. For 1,000, 5,000, and
10,000 mg/l units, ordinary commercial grade acids and persulfate are suitable. For the
KHP calibration standard, use reagent grade or better for all units.
• Rinse all glassware and plastic containers thoroughly with deionized water. To achieve the
best results, we recommend washing them once more with acid before use and rinsing
them thoroughly with deionized water.
• Use good scales and weigh the standard as accurately as possible before mixing. Keep
containers closed to prevent contamination and deterioration.
• Accurate preparation of the calibration standard is essential for accurate calibration of the
analyzer. Inaccurate preparation of the calibration standard will result in a faulty
calibration, which in turn will yield incorrect results.
Preparing the parent solution using potassium hydrogen phthalate
!
CAUTION
Potassium hydrogen phthalate (KHP) can irritate the skin and cause problems with the
eyes or respiratory channel.
‣ Do not inhale the powder and do not attempt to ever swallow any of the solution
prepared.
Shelf life: Storable parent solutions with concentrations of 5,000 mg/l are stable for
12 months if stored in a cool, dark place at 4 to 8 °C (40 to 46 °F). Prepared standard
solutions must be used within four weeks even if stored in a cool, dark place.
Concentrations: depends on the application.
16
Endress+Hauser
TOCII CA72TOC
Installation
Procedure:
For an organic carbon solution with a concentration of 5,000 mg/l, use a 1-liter volumetric
flask to dissolve 10,627 g of p.a. KHP in 500 - 700 ml deionized water (other concentrations
of the solution can be found in the following table). Once the chemical has dissolved, fill the
volumetric flask up to the mark. Stir the solution once more. Label the container, indicating
the contents and the date of preparation.
mg/l carbon
g KHP
Volume
5.000
10,627
1l
1.000
2,1254
1l
500
1,0627
1l
100
0,21254
1l
10
0,021254
1l
The solution with a concentration of 5,000 mg/l is stable (particularly when stored in a cool,
dark place). It is recommended to perform serial dilutions on the basis of a standard
substance of 5,000 mg/l to reduce the concentrations if needed. Example:
• If 10 ml of a solution with a concentration of 5,000 mg/l is diluted with 90 ml deionized
water, this results in a concentration of 500 mg/l.
• If 10 ml of a solution with a concentration of 500 mg/l is diluted with 90 ml deionized
water, this results in a concentration of 50 mg/l.
• If 10 ml of a solution with a concentration of 50 mg/l is diluted with 90 ml deionized
water, this results in a concentration of 5 mg/l.
Serial dilution is the preferred method for creating lower concentrations. This is instead of
diluting 1 ml of the 5,000 mg/l solution with 99 ml of deionized water. The latter method
can result in measurement errors.
NOTICE
Please note the following:
‣ Storable parent solutions with concentrations of 1,000 and 5,000 mg/l are relatively
stable for several weeks at room temperatures. Parent solutions must be stored in cool
dark place.
‣ In the case of a solution with a concentration of 10 mg/l, on the other hand, the quality
starts to deteriorate at room temperature within 3 to 5 days. For improved stabilization
of KHP standard solutions, nitric acid or sulfuric acid can be used for acidification
purposes. Use 4 ml of the 25% nitric acid (HNO3) or 4 ml of the 20% sulfuric acid (H2SO4)
per liter.
‣ If parent solutions with a high KHP content acidify, there is the risk of the KHP
precipitating.
‣ Make sure to keep the container with the crystalline KHP sealed at all times. If the
crystalline KHP comes into contact with the air, it absorbs water very quickly and must be
dried before use. Otherwise you will get inaccurate measurements since the
concentration of carbon is lower in the hydrous salt. KHP that has come into contact with
air must be dried for one hour at 105 °C (221 °F).
Endress+Hauser
17
Installation
TOCII CA72TOC
Preparing the parent solution from citric acid
!
WARNING
Nitric acid is highly caustic. Citric acid can irritate the skin and eyes and cause
respiratory problems.
‣ Wear protective goggles, gloves and protective clothing.
‣ Always add acids to water, not vice versa.
‣ Never attempt to swallow any of the solution prepared.
Shelf life: Storable parent solutions with concentrations of 100,000 mg/l are stable for
12 months if stored in a cool, dark place at 4 to 8 °C (40 to 46 °F). Prepared standard
solutions must be used within four weeks even if stored in a cool, dark place.
Concentrations: depends on the application.
Material:
• 1-liter volumetric flask
• Crtric acid monohydrate C6H8O7 · H2O (purity: p.a.)
• Nitric acid HNO3 (65 %, purity: p.a.)
Procedure:
For an organic carbon solution with a concentration of 100,000 mg/l, dissolve 291.6 g of
citric acid monohydrate in 500 ml of deionized water in a 1-liter volumetric flask (other
concentrations of the solution are listed in the table below). Carefully add 55.0 ml (77.0 g)
of nitric acid. Once the chemicals have dissolved, top up the mixture with water until the
1-liter mark is reached. Stir the solution once more. Label the container, indicating the
contents and the date of preparation.
mg/l carbon
Citric acid in g
Nitric acid in ml
Volume
100,000
291.6
55.0
1l
50,000
145.8
55.0
1l
10,000
29.16
55.0
1l
The solution with a concentration of 100,000 mg/l is stable (particularly when refrigerated).
To reduce the concentrations if necessary, serial dilutions based on a standard substance of
100,000 mg/l are recommended. Example:
• If 10 ml of a solution with a concentration of 100,000 mg/l is diluted with 90 ml deionized
water, this results in a concentration of 10,000 mg/l.
• If 10 ml of a solution with a concentration of 10,000 mg/l is diluted with 90 ml deionized
water, this results in a concentration of 1,000 mg/l.
• If 10 ml of a solution with a concentration of 1,000 mg/l is diluted with 90 ml deionized
water, this results in a concentration of 100 mg/l.
Serial dilution is the preferred method for creating lower concentrations. This is instead of
diluting 1 ml of the 100,000 mg/l solution with 99 ml deionized water. The latter method
can result in measurement errors.
Preparing the stripping reagent
Stripping reagent dosing is regulated via the pH sensor. The regulation range of the dosing
is from the minimum feed rate of the acid pump to approximately 300 times this amount.
The necessary acid quantity varies greatly from measuring place to measuring place. Ideally,
the strength of the acid in the feeder tank should be set in such a way that the system can
regulate in both directions, but the regulation range should be higher for higher volumes of
acid dosed.
18
Endress+Hauser
TOCII CA72TOC
Installation
Regulation aid:
The concentration of the acid is typically in the region of 5 to 10 %. The aim should be to
achieve a feed rate of 2 to 5 % (1.7 μl/min to 4.4 μl/min) for pump P3 (the current feed rate
can be read in the PROGRAMMING/TEST/OUTPUT/PUMPS menu).
For the acid feeder, mix 0.5 l deionized (DI) water with 0.125 l HNO3 25 % HNO3.
• Fill the acid hose
• Start measuring operation with a real sample
• Allow the acid dosing system to regulate
• Feed rate > 2 % < 5 %, note the acid concentration and use for future mixtures
• Feed rate < 2 % acid concentration, dilute preparation
• Feed rate > 5 % acid concentration, increase the concentration of the preparation
• Replace the contents of the acid hose
• Allow the acid dosing system to regulate, read off the feed rate
Increase concentration
Dilute
Deionized water (ml)
HNO3 25 % (ml)
HNO3 concentration
500
+ 125
5.0 %
+ 125
8.3 %
+ 125
10.7 %
+125
12.5 %
+125
5.0 %
500
+ 500
2.8 %
+1000
1.5 %
+2000
0.8 %
Use of other acids
• Nitric acid
Nitric acid does not form lyophobic salts in the strip chamber. The resulting nitrogen
oxides in the combustion gas are filtered out with an acid filter upstream from the IR
detector.
• Phosphoric acid
Phosphoric acid is a low-volatility, odorless acid that generates practically no precipitate.
Following combustion, substances are formed that are filtered out completely in the
heated salt trap.
• Sulfuric acid
Sulfuric acid is a low-volatility, odorless acid. When using sulfuric acid, precipitate can be
expected with alkaline earth metal ions. In addition to the sulfur dioxides, salts are
generated as a reaction product downstream from the furnace. These salts are easily
removed in the heated salt trap. Gypsum can also occur if the water is very hard.
• Hydrochloric acid
Hydrochloric acid is a volatile acid with a pungent odor. Salts do not precipitate in the strip
chamber. During combustion, salts occur (heatable salt trap) as does HCl gas, which is
absorbed in the condensate.
3.3
Post-installation check
• After installation, check that all connections are tightly fitted and are leak-resistant.
• Check all hoses for damage.
Endress+Hauser
19
Electrical connection
TOCII CA72TOC
4
Electrical connection
4.1
Wiring
!
WARNING
Device is live
Incorrect connection may result in injury or death.
‣ The electrical connection must only be carried out by a certified electrician.
‣ Technical personnel must have read and understood the instructions in this manual and
must adhere to them.
‣ Prior to commencing work on the connections, ensure that no voltage is applied to any
of the cables.
4.1.1
!
Preparing to wire
WARNING
Danger! Risk of electric shock!
‣ Disconnect the device from the power supply system (disconnect the mains plug). The
line filter, the overvoltage module and the main switch are still connected to the power
supply even when the main switch is switched off.
‣ Before connecting the device, make sure that the line voltage matches the voltage
indicated on the nameplate.
‣ Ensure that the analyzer is sufficiently grounded via the mains connection.
The analyzer is available for the following line voltages.
• 115 V AC 50 Hz
• 115 V AC 60 Hz
• 230 V AC 50 Hz
• 230 V AC 60 Hz
The following condition must be met when grounding the analyzer via the mains
connection:
50 V < R · Imax
Imax = maximum current at which the residual current circuit breaker is not yet triggered.
R = the resistance between protective ground and device ground.
If this condition cannot be guaranteed, the device must be grounded on site.
You must make the following connections:
• Alternating current connection via mains plug
• External grounding if necessary
• Analog 0/4 to 20 mA outputs
• Binary outputs
• Binary inputs
• RS-232
The signal connections are located in the EMC screening box on the right-hand side of the
cabinet.
The connection for the external ground is located at the bottom on the left-hand side of the
cabinet.
20
Endress+Hauser
TOCII CA72TOC
Electrical connection
4.1.2
Power distribution
a0025863
Fig. 5:
Endress+Hauser
Power distribution wiring diagram
21
Electrical connection
TOCII CA72TOC
Terminal strip assignment:
Connection
Description
3
Main switch, power distribution
4
Solenoid valve 3 - stripping gas
6
Peltier cooler regulator
7
Membrane compressor
8
Tube furnace
9
External salt trap
10
Solenoid valve 4, standard 1 + 2
11
Solenoid valve 7, carrier gas
13
Solenoid valve 5, bypass screen rinsing
14
Solenoid valve 1, sample/standard
15
Solenoid valve 6, channel 1/channel 2 (optional)
16
24 V power supply
17
Solenoid valve 2, strip chamber rinsing
18
Solenoid valve 8, dosing
Relay module assignment:
Relay No.
Relay type
Function
1
4A
Solenoid valve 1, sample/standard
2
3A
Solenoid valve 2, strip chamber rinsing
3
25A
Solenoid valve 3 stripping gas, tube furnace regulator, external salt trap
regulator, Peltier cooler regulator, membrane compressor
4
4A
Solenoid valve 4, standard 1 + 2
5
4A
Solenoid valve 5, bypass flush
6
4A
Solenoid valve 6 channel 1 / 2
7
4A
Solenoid valve 7, carrier gas
8
3A
Solenoid valve 8, dosing
RA
25A
Emergency shutdown
RB
25A
Heating regulator furnace
RC
25A
Heating salt trap
The power distribution is located at the back in the top door.
22
Endress+Hauser
TOCII CA72TOC
Electrical connection
4.1.3
Signal connection
I
1
1
II
2
1
2
0/4-20
mA
+
0/4-20
mA
+
binary in
2
III
1
IV
binary out
2
40
-
41
-
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
3
4
5
6
7
8
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
RS 232
a0011911
Fig. 6:
I
II
III
IV
40
41
Endress+Hauser
Signal connection
Alert
Alarm for limits
Standby
Operating control
Signal output channel 1
Signal output channel 2 (optional)
1
2
3
4
5
6
7
8
Calibration, external activation
Adjustment, external activation
Screen flush, external activation
Power flush, external activation
Not assigned
Not assigned
Standby, external activation
Channel switchover, external activation
(optional)
Messages I to IV
Potential-free relay contact (max. 0.2 A and 50 V), normally
closed (NC)
Relay contact I closed = no fault messages
Relay contact II closed = no collective alarm
Relay contact III closed = standby
Relay contact IV closed = operating check
At the end of a measuring cycle, relay IV opens for 2 seconds to
signal the end of the measuring cycle.
Signal outputs 40 to
41
0 to 20 mA or 4 to 20 mA can be switched, galvanically
isolated, max. load 500
Signal inputs 1 to 8
24 V DC active, max. load 500
23
Electrical connection
TOCII CA72TOC
Signal
input
Description
Switch status off
(open)
Switch status on
(closed)
1
Calibration, external activation
Analyzer is in measuring
mode
Calibration is triggered
2
Adjustment, external activation
Analyzer is in measuring
mode
Adjustment is triggered
3
Screen flush, external activation
Analyzer is in measuring
mode
Screen flush is triggered
4
Power flush, external activation
Analyzer is in measuring
mode
Power flush is triggered
5
Not assigned
6
Not assigned
7
Standby, external activation
Analyzer quits standby
mode and returns to
measuring mode or is in
measuring mode
Standby is triggered.
Analyzer is prepared for
standby. Standby is
maintained as long as the
switch status is closed.
8
Channel switchover, external
activation (optional)
Analyzer is in measuring
mode for the selected
channel.
Channel switch is
performed.
The potential-free contact must be closed for approx. 2 seconds to allow the switch
status to be triggered.
TOCII CA72TOC is suitable for connection to industrial power supply systems in
accordance with EN 61326-1, Class A.
24
Endress+Hauser
TOCII CA72TOC
Electrical connection
4.1.4
Power unit
24V +-
23A
24V +24V +24V +T
+
-
23
21
20
22
a0011956
Fig. 7:
Power unit assignment
Power unit wiring arrangement:
Connection
Description
20
Pump control 24 V DC
21
Magnetic stirrer control 24 V DC
22
Motor
23
Relay module 24 V DC
23A
Ventilator 24 V DC
The connections for the power unit are located on the rear side of the computer.
Endress+Hauser
25
Electrical connection
TOCII CA72TOC
4.1.5
Distributor connection
a0021331
Fig. 8:
Distributor (* = Pin 1 at MI1 - MI4 and at PWM)
Distributor assignment
Connection
Description
FI-24
NDIR detector
FI-26
pH - amplifier
BI-28
Carrier gas pressure switch DI 06
BI-29
Leak detector DI 05
BI-30
Standby internal DI 04
BI-34
Peltier cooler regulator DI 01 + 02
BI-35
Pressure switch dilution water DI 03
PWM-1
Furnace regulator (pin 1 black, Pin 2 blue)
PWM-2
Salt trap regulator (pin 3 brown, pin 4 gray)
BO-39
Relay module
Pu-38
Pump control
Ext. 55
External connection socket
MI1
Temperature sensor, furnace regulator, type K (pin 4 green, pin 6 white)
MI2
Temperature sensor, furnace monitoring, type J (pin 4 green, pin 6 white)
MI3
Temperature sensor, salt trap regulator, type J (pin 4 black, pin 6 white)
MI4
Pressure sensor (pin 1 VS brown, pin 3 signal + black, pin 4 signal – gray, pin 6 GND blue)
4.1.6
26
Fuses
Assembly
Fuses
Power distribution
2.5 A, slow-blow, design: fine-wire fuse 6.3 x 32
Relay
4 A per relay, slow-blow, design: TR5
Power unit
2 A, slow-blow, design: fine-wire fuse 5 x 20
Endress+Hauser
TOCII CA72TOC
Electrical connection
4.2
Post-connection check
Carry out the following checks after electrical connection:
Device status and specifications
Note
Is the exterior of the analyzer or cable undamaged?
Visual inspection
Electrical connection
Note
Does the supply voltage correspond to the data on the nameplate?
230 V AC 50/60 Hz
115 V AC 50/60 Hz
Are current outputs shielded and connected?
Are the mounted cables strain-relieved?
Are the cable types properly separated?
Run power supply and signal lines
separately over the entire travel
distance. Separate cable channels are
ideal.
Are the cables routed without loops or crossovers?
Are power supply and signal lines connected correctly according to
wiring diagram?
Are all screw terminals tightened?
Are all cable entries mounted, tightened and leak-resistant?
Endress+Hauser
27
Operation
TOCII CA72TOC
5
Operation
5.1
Operating elements
a0011861
Fig. 9:
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
28
Display and operating elements
Main switch
Display and operating elements
USB connection
Pump P2 sample analysis
Junction box
Compressor switch
Ventilator
Separation chamber
Dosing valve
Ventilator
Injection unit
Acid filter
Tube furnace with catalyst
Air outlet (filter pad)
Combined filter (water trap)
16
17
18
19
20
21
22
23
24
25
26
27
28
29
Heated salt trap (optional)
Circuit gas flowmeter
Dilution water pump P5 (optional)
Acid hose connector
MV4 calibration standard changeover C1/C2
MV1 wastewater/calibration standard
changeover
Valve for online sample/manual sample
Pump P3, acid dosing
Pump P4, sample - dilution (optional)
Mixing chamber (optional)
Pump P1, sample - strip chamber/condensate
extraction
Condensate hose connector
Strip chamber with pH electrode
Sample conditioning
Endress+Hauser
TOCII CA72TOC
Operation
5.2
Display and keypad
1
2
3
4
5
6
7
9
8
a0007996
Fig. 10:
1
2
3
4
5
6
7
8
9
Display and operating elements
Display, 16 rows a 40 characters
Keyboard
Arrow keys (cursors)
Function key"Measuring"
Function key "Service"
Function key "Programming"
Function key "Help"
Key "Enter"
"CLR" key
"Operation" key
Pressing the "Operation key" takes you back to the measuring mode.
The display shows the measured value trend of the last six hours in
graphic form.
"Service" key
Pressing the "Service key" takes you to the maintenance mode.
The menu contains the following items:
• Pumps
• Calibration
• Cleaning
• Filter
"Programming" key
Once you have pressed the "Programming key" and entered the four-digit
code (see code card supplied), you will be taken to the programming
menu.
The menu contains the following items:
• Setting
You can configure the measuring device here.
• Lists
You can visualize the reports and alarm messages on the display here.
• Test
You can use test programs to test the functions of the device here.
Use the help key [?] to obtain additional information on the current
date and the program version.
Arrow keys
The arrow keys help you position the cursor on the screen.
The "right" arrow key allows you enter negative values for certain
parameters. A minus sign appears once this key has been pressed.
Endress+Hauser
29
Operation
TOCII CA72TOC
"Enter" key
The "Enter key" has the following functions:
• You launch a menu item.
• You start a program item.
• You always confirm an entry with the "Enter key".
• When performing maintenance and service tasks, you confirm each
step with the "Enter key" on completion of the step.
"?" key
A short help text on the program item appears when you press the "?
key".
Press the "? key" a second time to exit the help text.
"4" key
Calls up the limit value list.
The current limit value violations are displayed.
"5" key
Calls up the fault list.
The current faults and alarms are displayed.
"6" key
Calls up the selected automatic service.
The selected service and the remaining time in seconds until the next
service are displayed.
"7" key
On devices with two sample flows, you can use the "7" key to toggle
between the values displayed on the screen for the two sample flows.
"." key (period key)
The period key indicates the current procedural step in measuring mode.
If you press the period key again, the following data are displayed:
• Temperature
• pH value
• Pressure in gas circuit
• Feed rate of pump P3
"CLR" key
You can display the following information on the screen with the "CLR
key":
• Device type
• Software program version
• Device options
30
Endress+Hauser
TOCII CA72TOC
Operation
5.3
Operation during measuring mode
The analyzer has three operating modes:
• Measuring mode
• Service mode
• Programming mode
The measuring mode is fully automated. Manual intervention on the part of the user is not
possible.
11:39:04
1000
1
500
2
6 00
5 00
7 00
8 00
IR = 334 ppm
5
9 00
10 00 11 00
234
3
TOC
mg/l
4
a0012443
Fig. 11:
1
2
3
4
5
Display during measuring operation
Time
Load curve of the past 6 hours
Timeframe
Measured value
Measured value of the IR detector
5.3.1
Recording mode
Pressing the U key when in measuring mode takes you to recording mode. Using the arrow
keys, scroll between the measured values recorded:
• V 1 day earlier
• W 1 day later
• U 2 hours earlier
• T 2 hours later
Once you have selected the required period, press the Enter key.
Recording time:
• 14 days for one-channel operation
• 7 days for two-channel operation
Use the operation key to quit recording mode.
Endress+Hauser
31
Operation
TOCII CA72TOC
5.3.2
Spot view
The spot view is activated in the recording mode by pressing the Enter key.
The following information is displayed:
• Load curve
• Measured value
• Date
• Time
A small cursor on the display indicates where you are currently in the load curve.
1000
1
500
5 00
6 00
RECORD
DATE 08/29/08
7 00
8 00
9 00
10 00 11 00
2
VALUE= 238 mg/l
TIME= 8:30
a0012449-en
Fig. 12:
1
2
V
W
Spot view (example)
Time mark on the load curve
Measured value of the selected time
1 hour earlier
1 hour later
U
T
2 minutes earlier
2 minutes later
You can switch the spot view (cursor) off with the Enter key.
32
Endress+Hauser
TOCII CA72TOC
Operation
5.4
Standby mode
The standby mode can be used to operate the analyzer at measuring points where the flow
of sample is interrupted at times.
The option is available in the one-channel version with a PA-2 or PA-3 sample conditioning
system.
4
3
5
2
1
a0013853
Fig. 13:
1
2
3
4
5
Standby switch
Adapter fitting for the pressure monitor
Pressure monitor ¼"
Signal connection
Protection cap
Bypass screen
Mode of operation
If the flow of sample is interrupted, the pressure monitor reports this status to the computer
via switch input DI 04.
This causes the device to respond as follows:
• All the pumps are stopped.
• Pump P2 is drained.
• The strip chamber is rinsed.
• The current output is set to 0.0 mA.
• The analyzer operates without displaying a measured value.
The device resumes measuring operation automatically as soon as sample flows again.
Endress+Hauser
33
Commissioning
TOCII CA72TOC
6
Commissioning
6.1
Function check
!
CAUTION
Avoid damages resulting from incorrect power supply and incorrect or poorly executed
hose connections.
‣ Check all connections for correctness. In particular, verify that all the hose connections
are securely positioned to ensure that no leaks occur.
‣ Verify that the line voltage matches the voltage specified on the nameplate.
6.2
Start
6.2.1
Commissioning sequence
Preparation:
1.
Install the pH sensor in the strip chamber and connect the sensor cable to the amplifier.
2.
Remove the transportation lock (cable ties) on the furnace unlocking device.
3.
Place the combustion pipe insert with the catalyst into the furnace (see "Maintenance"
section) and install the heated salt trap (optional).
4.
Mount the hose casettes (see the "Maintenance" section).
5.
Place the stripping reagent in the reagent tray beneath the measuring device, and put
the standards C1 and C2 (see "Preparing the chemicals" section) in the reagent bottle
holders provided on the left side panel.
Starting the analyzer:
34
1.
Switch on the analyzer. The furnace begins to heat up.
2.
In the programming mode, configure the operating parameters of the analyzer.
3.
Adjust the pH sensor (> CALIBRATION > ADJUSTMENT pH SENSOR).
4.
Adjust the peristaltic pumps P1 and P4 (> PUMPS > REPLACE HOSE P1/4).
5.
Adjust the peristaltic pump P2 and determine the empty volume ( > PUMPS > ADJUST
P2 and > CALIBRATION > EMPTY VOLUME DOSING).
6.
Once the analyzer is in operation following warm-up and the temperature is stable,
check the gas circuit for leaks (> CLEANING > LEAKAGE TEST).
7.
Perform a 2-point adjustment ( > CALIBRATION > ANALYZER ADJUSTMENT).
Endress+Hauser
TOCII CA72TOC
Commissioning
6.2.2
Updating the analyzer software
Regardless of the menu structure, you can update the analyzer software via the USB
interface.
!
WARNING
Risk of electric shock due to defective device connected via USB!
‣ Use only passive USB devices.
Step
1
Display
Description
1. Switch off the main switch.
2. Plug the USB stick with the
desired software into the USB
interface.
3. Switch the main switch back on.
Logo appears.
2
Press the [CLR] key.
3
Press [1] key.
Keys [2] and [3] are reserved for E+H
Service.
[CLR] Cancel and start existing
analyzer software
4
A list of all software versions
available is displayed.
[▴▾] Scroll up and down
[◂▸] Page through (if more than 12
versions are available)
[.] Select software version (* =
marking)
[CLR] Delete software version (! =
marking)
[E] Confirm
Only one version can be
selected for updating. Several
versions can be selected for
deletion.
Endress+Hauser
35
Commissioning
TOCII CA72TOC
Step
5
Display
Description
[E] Confirm the selected software
version and start the software.
[CLR] Back to selection menu
The analyzer goes into
measuring mode once the
software has been started. You
can check the software version
when in measuring mode using
the [CLR] key.
6
Remove the USB stick once the
software has been updated.
The software versions will remain available in memory until they are deleted. For
improved clarity, it can be useful to delete these software versions if further updates are
carried out.
36
Endress+Hauser
TOCII CA72TOC
Commissioning
6.3
Programming mode
You configure the operating parameters of the analyzer in the programming mode.
1.
Press the programming key.
2.
Enter the "code". The code is a four-digit number which you can find on the code card
supplied with your analyzer.
3.
Press the [E] key.
The following menu appears:
a0012583-en
Fig. 14:
"Programming" menu
The operating parameters are stored in the analyzer and are retained in the event of a
software update.
Within the menu items, pressing the [<] key takes you back to the "Programming" menu.
6.3.1
Menu SETTING - RANGE DATA
a0012553-en
Fig. 15:
Endress+Hauser
Menu "Setting"
37
Commissioning
TOCII CA72TOC
Parameter
Unit
Factory
setting
Description
SCALE
mg TOC/l
1000
Use this function to enter the maximum concentration
for your measuring point here. This value determines
the scale end value for the graphic interface.
Enter two independent values for the two-channel
version.
SCREEN FLUSH
n/day
0
The number of automatic flush cycles for the bypass
screen per day (recommended value: 2).
DURA.SCREEN
FLUSH
s
15
When a screen flush is activated, the duration of the
flush can be altered. If longer than 15 seconds, two
thirds of the flush time is assigned to flushing the
screen and one third to flushing the strip chamber.
POWER FLUSH
n/day
0
The number of automatic power flush cycles for the
stripping vessel and separation chamber per day
(recommended value: 2).
PAUSE CYCLE
s
0
Interval between 2 measurements.
P1 (B)
ml/min
7.5
Feed rate of pump P1.
P2 (B)
μl/min
250
Feed rate of pump P2.
P4 (B)
ml/min
5.0
Feed rate of pump P4 (optional). The feed rates of
pumps P4 and P5 determine the dilution ratio.
P5 (B)
ml/min
25.0
Feed rate of pump P5 (optional). The feed rates of
pumps P4 and P5 determine the dilution ratio.
VOL.BATCH
μl
300
Dosing volume for a batch. Increasing the volume
increases the sensitivity of the measuring system but
also increases the salt load.
STANDARD C1
mg/l
80
Concentration of standard solution C1.
STANDARD C2
mg/l
720
Concentration of standard solution C2.
CAL./ADJUST.
n days
3
Here you can specify after how many days a calibration
or adjustment should be performed. The automatic
function is switched off with the value 0.
CAL./ADJUSTMENT
TIME
xx.xx
23.00
Here you can specify the start time for the calibration or
adjustment. Enter as a decimal number. Example:
22.50 corresponds to 22:30 hrs.
2
Use this function to specify which function should be
executed.
CAL/ADJUSTMENT
• 1 - Calibration
• 2 - Adjustment
The function is executed 90 minutes before the day
changes.
38
Endress+Hauser
TOCII CA72TOC
Commissioning
6.3.2
Menu SETTING - BASIC DATA
a0012580-en
Fig. 16:
"Programming" menu
Parameter
Unit
Factory
setting
Description
DC OUT
mA
4
Sets the signal output to 0 to 20 mA or 4 to 20 mA.
SCALE AO
mg/l
1000
Scale end value of the analog output.
e.g. 1000 mg/l = 20 mA
EMPTY VOLUME P2
*
μl
220
Dead volume of pump P2 from the separation chamber
to the end of the capillary.
P1 100% *
ml/min
5.6
Feed rate of pump P1 at 100 % pump capacity.
P2 100% *
μl/min
870
Feed rate of pump P2 at 100 % pump capacity.
P3 100%
μl/min
870
Feed rate of pump P3 at 100 % pump capacity.
P4 100% *
ml/min
5.6
Feed rate of pump P4 at 100 % pump capacity
(optional).
P5 100% *
ml/min
30
Feed rate of pump P5 at 100 % pump capacity
(optional).
ADJUSTMENT
CONSTANTS
Do not change this value!
X0 *
0.0
Offset, value is overwritten during adjustment
KP *
50
Slope, value is overwritten during adjustment
PH-CONTROL
1.00
The measuring device is equipped with automatic pH
control in the stripping vessel. You can use this
parameter to switch pH control on or off.
• 1.00 = pH control is switched on
Reading on display = TOC
• 0.00 = pH control is switched off
Reading on display = TC
Endress+Hauser
PH-NOMINAL
2.5
The pH value in the stripping vessel is set here. The pH
value must be between 1 and 4 for complete stripping.
If the sample becomes excessively acidic in municipal
sewage treatment plants, there is the problem of humic
acid precipitating which could mask the carbonate. This
inorganic carbon enters the furnace and results in
excessively high readings.
PH ADJ.OFFSET *
2.4
Offset of the pH sensor. The value is overwritten when
the pH sensor is adjusted.
39
Commissioning
TOCII CA72TOC
Parameter
Unit
Factory
setting
Description
PH ADJ.SLOPE *
mV/
decade
57.5
Slope of the pH sensor. The value is overwritten when
the pH sensor is adjusted.
* These parameters are adapted by means of menu-guided adjustments.
6.3.3
Menu SETTING - ALARM LIMITS
Use this function to enter the limit values for the high alarm and low alarm.
a0012579-en
Fig. 17:
40
"Programming" menu
Parameter
Unit
Factory
setting
Description
HIGH ALARM LIMIT
mg/l
12000
Use this function to specify the alarm value for when
the value is overshot.
LOW ALARM LIMIT
mg/l
0
Use this function to specify the alarm value for when
the value is undershot.
Endress+Hauser
TOCII CA72TOC
Commissioning
6.3.4
Menu SETTING - SET CLOCK
This section describes how you enter the date and the time.
Select the programming mode for this purpose.
Step
Display
Description
1
2
[] increases value by 1.
[] reduces value by 1.
[] to go to previous value
[] to go to next value
[E] confirms the displayed value.
3
6.3.5
Step
Menu SETTING - SET BRIGHTN./CONTR.
Display
Description
1
2
Endress+Hauser
[E] confirms the displayed value.
[] and [] switch between
brightness and contrast.
[] increases value by 1.
Use [] reduces value by 1.
Range: 0 to 100 %
41
Commissioning
TOCII CA72TOC
6.3.6
Menu SETTING - MEASURING SITE
You can give the measuring point a name.
a0012582-en
Fig. 18:
"Programming" menu
In the factory setting, the name is MEASURING SITE.
• Select the position of the character using the cursor buttons T and U.
• Select the character using the V and W cursor buttons.
• Pressing the "1" key switches the character to "A".
6.3.7
Menu LISTS - SHOW COMPLETE RECORDS
a0012588-en
Fig. 19:
"Lists" menu
Chronological display of all the events saved. The last 200 events are saved in the complete
record list.
XX.XX
XX.XX
Date
XX:XX
XX:XX
Time
XXXXXXXXXXXXX
XXXXXXXXXXXXX
Event
a0012593-en
Fig. 20:
42
Complete records
Endress+Hauser
TOCII CA72TOC
Commissioning
6.3.8
Menu LISTS - ALARM RECORD
a0012587-en
Fig. 21:
"Lists" menu
All the alarm messages, including the date and time of the event, are logged in the alarm
record.
ALARM RECORD
Date
XX.XX
XX.XX
Time
XX:XX
XX:XX
Event
on/off
XXXXXXXXXXXXX XXX
XXXXXXXXXXXXX XXX
a0012592-en
Fig. 22:
Endress+Hauser
Alarm record
Alarm
Description
ALARM TMEASURING RANGE
The value is outside the measuring range
The IR detector has been working above its maximum value for over 10
minutes or the measuring device has been measuring 0 mg/l for over an
hour.
MALFUNCTION IR
IR detector is defective
Operation stops. The analyzer restarts automatically once the error
status is reset.
LEAKAGE
Leak in the system
The furnace and carrier gas supply will be switched off.
The analyzer must be started manually.
43
Commissioning
44
TOCII CA72TOC
Alarm
Description
ADJUSTMENT ERROR
Adjustment error
An error number is assigned to the error.
UNSTABLE DOSING
Error when dosing the sample
The minimum number of drops to be expected was undershot.
WATER PRESS. FAILURE
Failure in the supply of water for flushing and dilution
The pressure is below the minimum permitted pressure of approx. 1.5
bar (21 psi).
Operation stops. The analyzer restarts automatically once the error
status is reset.
CHECK CO2 BASIC LINE
The CO2 threshold value (ppm) or the CO2 drift (ppm/min) was
overshot in basic line measurement
INPUT ERROR C1>C2
Error during entry of standard concentrations. The concentration of
standard C1 must be lower than the concentration of standard C2.
CIRCUIT PRESSURE HIGH
At 175 mbar, the pressure in the gas circuit is 70 % above the pressure
permitted in the gas circuit (250 mbar).
CIRCUIT PRES.TOO HIGH
The maximum permitted pressure in the gas circuit has been exceeded.
The max. set value for pressure is 250 mbar.
INTERNAL COM-FAULT
Fault in internal communication between I/O card, keyboard and
modbus connection. Operation stops.
The analyzer restarts automatically once the error status is reset.
Endress+Hauser
TOCII CA72TOC
Commissioning
6.3.9
Menu LISTS - MAINTENANCE RECORD
a0012586-en
Fig. 23:
"Lists" menu
All the maintenance procedures are recorded by maintenance activity in the maintenance
record. Maintenance procedures that have not been carried out cannot be selected.
Endress+Hauser
Action
Description
PROGRAM STARTED
Date and time when program starts
CHANGE DATA
Date and time of when the setting data are changed
CHANGE TIME
Date and time when clock changes
The newly set time and the time difference in hours compared to the
old time are documented.
Negative value: The clock was put back.
Positive value: The clock was put forward.
ADJUSTMENT
Date and time of when the analyzer and the CO2 concentrations of the
standard solutions are adjusted
Value 1: CO2 concentration of C1 [ppm]
Value 2: CO2 concentration of C2 [ppm]
ADJUSTMENT CONSTANTS
Date and time and the adjustment constants obtained during
adjustment
Value 1: Offset [ppm]
Value 2: Standardized slope [ppm]
CALIBRATION
Date and time of the calibration of the analyzer and the TOC value
found, and the recovery with regard to the specified concentration of
standard C2
Value 1: TOC [mg/l]
Value 2: Recovery [%]
BASIC LINE DRIFT
Date and time of the basic line drift when calibrating and adjusting
Value 1: Basic line offset [ppm]
Value 2: Basic line drift slope [ppm/min]
EMPTY VOLUME DOSING
Date and time when selected in the service menu
Value 1: Duration of the filling process [s]
Value 2: Volume [μl]
ADJUSTMENT PUMP P1
Date and time of when pump P1 is adjusted
Value 1: new feed rate (ml/min)
Value 2: old feed rate (ml/min)
ADJUSTMENT PUMP P2
Date and time of when pump P2 is adjusted
Value 1: new feed rate (μl/min)
Value 2: old feed rate (μl/min)
ADJUSTMENT PUMP P4
Date and time of when pump P4 is adjusted
Value 1: new feed rate (ml/min)
Value 2: old feed rate (ml/min)
45
Commissioning
46
TOCII CA72TOC
Action
Description
ADJUSTMENT PH-SENSOR
Date and time and the adjustment constants obtained during
adjustment
Value 1: Offset [mV]
Value 2: Slope [mV/log pH]
REPLACE HOSE P1
Date and time of when the hose of pump P1 is changed
REPLACE HOSE P2
Date and time of when the hose of pump P2 is changed
REPLACE HOSE P3
Date and time of when the hose of pump P3 is changed
REPLACE HOSE P4
Date and time when the hose of pump P4 is changed when sample predilution is present
SCREEN FLUSH
Date and time when selected in the service menu
Automatic screen flushing is not logged.
BYPASS SCREEN
Date and time when selected in the service menu
POWER FLUSH
Date and time when selected in the service menu
Automatic power flushing is not logged.
STRIPPING+SEPARATION
Date and time when selected in the service menu
OPEN GAS CIRCUIT
Date and time when selected in the service menu
COMBUSTION PIPE
Date and time when selected in the service menu
LEAKAGE TEST
Date and time when exiting the leak tightness display
Value 1: Current pressure
Value 2: Current leakage rate [mbar/min]
Typical value: -0.5 to -2.0 mbar/min
REPLACE ACID FILTER
Date and time when selected in the service menu
REPLACE GAS FILTERS
Date and time when selected in the service menu
HEATED FILTER
Date and time when selected in the service menu
(heated salt trap)
REPLACE O2 FILTER
Date and time when selected in the service menu
STANDBY
Date and time of a standby event
SAVE DEFAULTS
Date and time when selected in the Setting menu
SET DEFAULTS
Date and time when selected in the Setting menu
Endress+Hauser
TOCII CA72TOC
Commissioning
6.3.10
Menu LISTS - MIN MAX AVERAGE
a0012585-en
Fig. 24:
"Lists" menu
This menu item displays the maximum, minimum and average measured values for the days
saved.
6.3.11
Menu LISTS - RECORD ALL DATA
This menu item is used to save the measuring data and protocols from the last 14 days. The
data records are available as csv files.
Step
Display
Description
1
2
RECORD DATA
01
02
03
04
05
06
07
-
06.02.2013
05.02.2013
04.02.2013
03.02.2013
02.02.2013
01.02.2013
31.01.2013
[E] TO END
3
Endress+Hauser
08
09
10
11
12
13
-
30.01.2013
29.01.2013
28.01.2013
27.01.2013
26.01.2013
25.01.2013
If the time or the date are changed
within these 14 days, the data will be
re-dated accordingly.
If the date change is outside of these
14 days, the memory will be
completely deleted.
[1] SAVE
Connect the storage medium to the
USB connection.
47
Commissioning
TOCII CA72TOC
Step
4
Display
Description
RECORD DATA
01
02
03
04
05
06
07
-
06.02.2013
05.02.2013
04.02.2013
03.02.2013
02.02.2013
01.02.2013
31.01.2013
08
09
10
11
12
13
SAVE DATA, PLEASE WAIT
5
USB CAN NOW BE REMOVED
-
30.01.2013
29.01.2013
28.01.2013
27.01.2013
26.01.2013
25.01.2013
01:55
Always remove the storage medium!
[E] TO END
48
Endress+Hauser
TOCII CA72TOC
Commissioning
6.3.12
Step
Menu TEST - INPUT TEST
Display
Description
1
1. Select the desired inputs.
2. Press the [E] key.
2
Analog inputs
Current CO2 measured value
• T1 = Temperature furnace
monitoring
• T2 = Temperature heating
regulation furnace, PWM
performance display
• T3 = Temperature heating
regulation salt trap, PWM
performance display
• pH value in stripping vessel
• Pressure level in gas circuit
Switch inputs
Switching status of the switch inputs
Ix = 0 = OFF
Ix > 0 = ON
IN1= Peltier cooler, Peltier regulator
BI34
IN2= Peltier cooler, Peltier regulator
IN3= Dilution water BI35
IN4= Standby no. BI30
IN5= Leak detector no. BI29
IN6= Carrier gas pressure switch BI28
Endress+Hauser
49
Commissioning
TOCII CA72TOC
6.3.13
Step
Menu TEST - OUTPUT TEST
Display
Description
1
1. Select the desired outputs.
2. Press the [E] key.
2
3
MEAUSREMENT OFF deactivates measuring mode. The MEASUREMENT OFF operating status is
displayed. Select this status to test the inputs without triggering an alarm.
TEST
DC-SIGNAL
CHANNEL NO. 1
CHANNEL NO. 2
4
Pumps
Parameter for testing the function of
the pumps
Negative value turns the direction of
flow.
5
Switch outputs
Displays the switching status of the
switch outputs (see the following
table).
ON/OFF using Enter key
6
TEST COM
...
[E] TO END
50
0.0 mA
0.0 mA
Current output
Sets the analog current outputs to any
value between 0 and 20 mA.
TEST COM displays the transmission
data for the computer interface
RS 232.
The menu item allows users to test
the transmission of data with an
external terminal.
Once the data connection is
established, a data string is
transmitted every 2 seconds.
Keystrokes at the external terminal
are shown on the display. Entries at
the terminal must be transmitted
using a carriage return.
Endress+Hauser
TOCII CA72TOC
Commissioning
Output
Description
Switch status OFF
(contacts opened)
Switch status ON
(contacts closed)
SA1
Changeover between standard and
sample
Sample
Standard solution
SA2
Flushing valve for power flushing
Strip chamber flushing
switched off
Strip chamber flushing
switched on
SA3
Carrier gas supply, tube furnace
regulator, Peltier cooler regulator,
membrane compressor
Consumer load switched off
Switch status during
measuring operation
SA4
Changeover between standard 1 and
standard 2
Standard 1
Standard 2
SA5
Screen flushing valve
Screen flushing valve off
Screen flushing valve on
SA6
Optional: Changeover between channel
1 and channel 2
Channel 1
Channel 2
SA7
Carrier gas flushing valve
Carrier gas flushing off
Carrier gas flushing on
SA8
Dosing valve
Dosing valve open
Dosing valve closed
SA9
Collective alarm fault relay I
(e.g. failure acid, leak)
Fault on
Fault off
SA10
Collective alarm limit values relay II
Limit value alarm on
Limit value alarm off
SA11
Standby relay III
Standby off
Standby on
SA12
Operation check relay IV
At the end of the measuring
cycle in measuring mode,
the contact is opened for 2
sec to signal the end of the
measuring cycle. If the
analyzer is in service mode
or in a fault state in which
TOC measurement is not
permitted, the contact is
opened.
During operation, the
contact is closed as soon as
the displayed measured
value is reliable (e.g.
following a service, this
contact is closed once the
first TOC measured value
has been determined).
6.3.14
Two-channel: external changeover
The analyzer is equipped with one or two separate sample supply systems.
The current sample selected is controlled externally by signal input 8 (binary in 8).
Signal input 8 = 0 channel 1
Signal input 8 = 1 channel 2
If the analyzer is operated with one sample conditioning system, the operator must ensure
that the right sample is at the bypass when a channel switch is requested.
If the analyzer is operated with two sample conditioning systems, the solenoid valve MV6 is
used internally to switch the channels.
If the signal status at signal input 8 changes, the measuring cycle is terminated immediately
and the system immediately commences channel switchover.
If the "Operation" key is pressed during channel switchover, the process of switching
channels is aborted and the measuring cycle commences in the active channel. Conditioning
of the analyzer to the sample in the active channel is suppressed.
The measuring channel cannot be switched manually.
Endress+Hauser
51
Commissioning
TOCII CA72TOC
Settings for the graphic interface
You must enter the following parameters in the Range Data menu:
PROGRAMMING/SETTING/RANGE DATA:
Parameter
Unit
Description
SCALE CH1
mg TOC/l
Use this function to enter the maximum concentration for
channel 1 here. This value determines the scale end value for
the graphic interface.
SCALE CH2
mg TOC/l
Use this function to enter the maximum concentration for
channel 2 here. This value determines the scale end value for
the graphic interface.
Settings for the analog outputs
You must enter the following parameters in the Basic Data menu:
PROGRAMMING/SETTING/BASIC DATA:
Parameter
Unit
Description
SCALE AO CH1
Scale end value of the analog output of channel 1.
SCALE AO CH2
Scale end value of the analog output of channel 2.
Settings for the alarm limits
You must enter the following parameters in the ALARM LIMITS menu:
PROGRAMMING/SETTING/ALARM LIMITS:
Parameter
Unit
Description
HI ALARM LIMIT CH1
mg/l
Use this function to specify the alarm value for when the value
for channel 1 is overshot.
LO ALARM LIMIT CH1
mg/l
Use this function to specify the alarm value for when the value
for channel 1 is undershot.
HI ALARM LIMIT CH2
mg/l
Use this function to specify the alarm value for when the value
for channel 2 is overshot.
LO ALARM LIMIT CH2
mg/l
Use this function to specify the alarm value for when the value
for channel 2 is undershot.
All the alarm limits affect the same signal output II (binary out II). A limit value alarm is also
retained following channel switchover until the limit value for the channel in question is
undershot.
52
Endress+Hauser
TOCII CA72TOC
Commissioning
6.3.15
Two-channel: time-controlled changeover
The analyzer is equipped with two separate sample supply systems.
Configuring the measuring duration
The measuring duration can be individually configured for each channel.
PROGRAMMING/SETTING/BASIC DATA:
Parameter
Unit
Description
DURATION CH1
min
Measuring duration for channel 1
DURATION CH2
min
Measuring duration for channel 2
If you configure a duration of 0 minutes in one channel, measurement will be performed
permanently in the other channel. You have to set a duration greater than 0 minutes for at
least one channel.
Regardless of the measuring duration configured, any measuring cycle that has started
will always be completed before the system switches to the other channel.
In the case of an external switchover, the other channel is switched over to immediately.
Settings for the graphic interface
Using key 7 you can switch over to the graphic screen for the channel which was not
selected.
You must enter the following parameters in the Range Data menu:
PROGRAMMING/SETTING/RANGE DATA:
Parameter
Unit
Description
SCALE CH1
mg TOC/l
Use this function to enter the maximum concentration for
channel 1 here. This value determines the scale end value for
the graphic interface.
SCALE CH2
mg TOC/l
Use this function to enter the maximum concentration for
channel 2 here. This value determines the scale end value for
the graphic interface.
Settings for the analog outputs
You must enter the following parameters in the Basic Data menu:
PROGRAMMING/SETTING/BASIC DATA:
Parameter
Endress+Hauser
Unit
Description
SCALE AO CH1
Scale end value of the analog output of channel 1.
SCALE AO CH2
Scale end value of the analog output of channel 2.
53
Commissioning
TOCII CA72TOC
Settings for the alarm limits
You can enter the following parameters in the ALARM LIMITS menu:
PROGRAMMING/SETTING/ALARM LIMITS:
Parameter
Unit
Description
HI ALARM LIMIT CH1
mg/l
Use this function to specify the alarm value for when the value
for channel 1 is overshot.
LO ALARM LIMIT CH1
mg/l
Use this function to specify the alarm value for when the value
for channel 1 is undershot.
HI ALARM LIMIT CH2
mg/l
Use this function to specify the alarm value for when the value
for channel 2 is overshot.
LO ALARM LIMIT CH2
mg/l
Use this function to specify the alarm value for when the value
for channel 2 is undershot.
All the alarm limits affect the same signal output II (binary out II). A limit value alarm is also
retained following channel switchover until the limit value for the channel in question is
undershot.
Interrupting the time control system
Regardless of the internal time-based control system, the channel can be changed manually,
or by remote control via the external signal input 8 (binary in 8).
Using keys 1 and 2 users can manually change to the other channel during measuring
operation.
Channel changeover can be triggered remotely using signal input 8.
Signal input 8 = 0 has no effect
Signal input 8 = 1 (for approx. 10 seconds) changes over to another channel
If you trigger channel changeover by using the keyboard or the signal input, the measuring
cycle is terminated immediately and the channel changeover initiated.
54
Endress+Hauser
TOCII CA72TOC
Commissioning
6.4
Optimizing the analyzer
6.4.1
Adjusting the measuring range
Depending on its configuration, the analyzer can measure from just a few mg/l to several
10 000 mg/l. The analyzer can be optimized by changing components and altering the
device settings.
The measuring range is altered by changing a component:
• By changing the infrared detector (Endress+Hauser Service)
• By fitting a predilution system (Endress+Hauser Service)
In terms of device settings, the feed rate of dosing pump P2 is optimized.
Please note that optimizing the sensitivity or the salt load compatibility often demands
conflicting settings on the analyzer. Select the settings the offer the best compromise for
your measuring task.
6.4.2
Influence of dosing volume (pump P2) on the measuring signal
The dosing range of adjustment is affected by:
• Dosing volume BATCH VOL. (100 to 1200 μl)
You can change the pump setting in the programming mode: SETTING - RANGE DATA.
An increase in dosing volume (P2) increases the measuring signal, with a 50 % increase in
the feed rate being equivalent to an approx. 50 % increase in the signal.
The dosing volume of 400 μl/batch results in a sensitivity of 2 ppm CO2 per mg/l TOC
(dosing volume 100 to 1200 μl/batch is equivalent to 0.5 to 6 ppm CO2 per mg/l TOC).
Please note that if the sample flow is increased, the salt load also increases to the same
extent.
The maximum measuring range indicated on the nameplate is the range at a dosing volume
of 100 μl/batch and 1200 μl/batch.
Measuring range
Dosing
Set measuring range
0.25 to 600 mg TOC/l
100 μl/batch
300 μl/batch
1200 μl/batch
3 to 600 mg/l
1 to 200 mg/l
0.25 to 50 mg/l
1 to 2400 mg TOC/l
100 μl/batch
300 μl/batch
1200 μl/batch
12 to 2400 mg/l
4 to 800 mg/l
1 to 200 mg/l
2.5 to 6000 mg TOC/l
100 μl/batch
300 μl/batch
1200 μl/batch
20 to 6000 mg/l
8 to 2400 mg/l
2.5 to 500 mg/l
5 to 12000 mg TOC/l
100 μl/batch
300 μl/batch
1200 μl/batch
60 to 12000 mg/l
24 to 4800 mg/l
5 to 1000 mg/l
Factory setting:
Version A and B: 300 μl/batch
Version C and D: 250 μl/batch
Endress+Hauser
55
Commissioning
TOCII CA72TOC
6.4.3
Operating the analyzer with high salt loads
Completely diverse salt loads can occur given the different range of applications of the
analyzer. For the user, this can mean that the salt load reduction function has to be
optimized. The dosing pump P2 can be used to reduce the dosing volume and therefore the
salt load. To reduce the salt load, it is also possible to program a break in measurement.
For very high salt loads, an optional dilution module is available.
Dilution ratios between 1:5 to 1:20 are possible. The effective TOC concentration in the
diluted wastewater should be inside the measuring range of the analyzer.
56
Endress+Hauser
TOCII CA72TOC
Commissioning
6.5
Communication
The analyzer is equipped with an RS-232 serial interface.
Data transmission is unidirectional and is performed with the following parameters:
Baud rate
9600 baud
Bits
8 bit
Parity
N
Stop bit
1 bit
Handshake
No
The string is 104 bytes long and is sent every 2 seconds.
Endress+Hauser
Byte
Description
0
Start byte
1
0 = Operation disabled
1 = Operation enabled
2
0 = Emergency shutdown
1 = Channel 1 measuring operation enabled
2 = Adjustment or calibration
3 = Service
4 = Programming
5 = Channel 2 measuring operation enabled
3
Leakage (0 = off, 1 = on)
4
Temperature too high (0 = off, 1 = on)
5
Low carrier gas supply (0 = off, 1 = on)
6
IR detector fault (0 = off, 1 = on)
7
Temperature too low (< 85 % Tset) (0 = off, 1 = on)
8
Outside the measuring range (0 = off, 1 = on)
9
Temperature deviation of Peltier cooler (Tset ± 3 °C) (0 = off, 1 = on)
10
pH alarm (0 = off, 1 = on)
11
Temperature deviation (< Tset -30 °C) (0 = off, 1 = on)
12
Standby (0 = off, 1 = on)
13
Limit value exceeded (0 = off, 1 = on)
14
Limit value undershot (0 = off, 1 = on)
15
Slope alarm (0 = off, 1 = on)
16
Unstable dosing, sample failure (furnace) (0 = off, 1 = on)
17
Water supply failure (0 = off, 1 = on)
57
Commissioning
58
TOCII CA72TOC
Byte
Description
18
Gas circuit pressure monitoring
0 = OK
1 = 70 % of max. permitted pressure
2 = > max. permitted pressure
19
Check CO2 basic line (0 = off, 1 = on)
20
Adjustment error (0 = off, 1 = on)
21
Empty (unused byte)
22
Empty (unused byte)
23
0 = No valid measured value available
1 = Valid measured value available
2 = New measured value determined (displayed for approx. 4 seconds)
24
Separator
25
0 = Sample
1 = Standard is dosed
26
Flushing strip and separation chamber with supply water
27
0 = Error shutdown active, no power is supplied to any of the units
powered by the power relay
1 = Power supply active
28
0 = Standard C1 is dosed
1 = Standard C2 is dosed
If relay 1 (byte 25) is set to 1
29
Flushing sample conditioning
30
Only important for two-channel operation
0 = Sample is taken from sample channel 1
1 = Sample is taken from sample channel 2
31
Flushing with carrier gas
32
A 0-1-0 change indicates that the process of dosing the sample into the
furnace has been completed.
33
Separator
34 ... 39
TOC measured value (mg/l)
1 decimal place for measuring range A and B
0 decimal places for measuring range C and D
40
Separator
41 ... 46
Only for channel 2
TOC measured value (mg/l)
1 decimal place for measuring range A and B
0 decimal places for measuring range C and D
Endress+Hauser
TOCII CA72TOC
Endress+Hauser
Commissioning
Byte
Description
47
Separator
48 ... 53
CO2 (ppm)
1 decimal place; current value of the gas card
54
Separator
55 ... 60
CO2 (ppm)
1 decimal place; CO2 difference calculated from the measuring cycle
61
Separator
62 ... 67
pH value
2 decimal places
68
Separator
69 ... 74
Number of drops dosed into furnace
No decimal places
75
Separator
76 ... 81
Batch status
82
Separator
83 ... 92
Date DD.MM.YYYY
93
Separator
94 ... 101
Time HH:MM:SS
102
Carriage return
103
Line feed
104
End of transmission
59
Diagnostics and troubleshooting
TOCII CA72TOC
7
Diagnostics and troubleshooting
The analyzer permanently monitors its functions automatically. Any errors detected by the
device are shown on the display.
!
WARNING
The device is energized.
Troubleshooting that is not performed correctly may result in injury or death.
‣ Troubleshooting work on the components behind the mounting plate may only be carried
out by an electrical technician.
!
CAUTION
There is a risk of infection when working with wastewater, and there is a risk of injury
when working with reagents.
‣ Wear acid-resistant protective gloves, protective goggles and a laboratory coat!
7.1
Error messages
Message
Cause
Tests or remedial action
Temperature
too high
The temperature at
the tube furnace is
70 °C above the set
point.
• Select INPUT TEST. The temperatures are displayed. If there
is a significant variation in the temperatures, check the
temperature sensor.
• It is possible that the furnace will be permanently heated.
Remove the PWM connection (cable 54). If the temperature
continues to rise, check relay RB.
• Select INPUT TEST. If the PWM control system continuously
outputs 200 %, there is a malfunction in the PWM Switch the
main switch off and on - if the error persists, replace the I/O
card.
Possible defect:
Temperature sensor,
relay RB, PWM1,
I/O card
Temperature
too low
The temperature is
15 % below the set
point.
When putting back
into operation, or
during service or
maintenance
Possible defect:
Cable, temperature
sensor, relay RB,
PWM1, I/O card
Temperature
below xxx °C
• Select INPUT TEST. The temperature are displayed. If the
temperature increases in INPUT TEST, wait until the warmup stage has stabilized. If there is a significant variation in the
temperatures, check the temperature sensor. Verify that the
temperature sensors are secured and positioned correctly on
the furnace pipe. Are the contacts on the furnace connection
socket and the I/O card correctly connected.
• It is possible that the furnace will not be heated. Check the
relay RB.
• Select INPUT TEST. If the PWM control system does not
regulate and continuously outputs 200 % or 0 %, there is a
malfunction in the PWM. Switch the main switch off and on if the error persists, replace the I/O card.
The measured
temperature is
30 °C below the set
temperature.
When putting back
into operation, or
during service or
maintenance
Possible defect:
Temperature sensor,
relay RB, PWM1,
I/O card
60
Endress+Hauser
TOCII CA72TOC
Diagnostics and troubleshooting
Message
Cause
Tests or remedial action
Carrier gas
failure
The pressure sensor
for monitoring the
carrier gas has been
triggered.
Pressure < 1.5 bar.
Failure of carrier gas
supply
• Monitor the carrier gas supply.
• Check signal processing (I/O card slot no. 28, switch input
DI06)
Go to the programming module PRG/TEST/INPUTS/SWITCH
INPUTS. Release the connecting cable on the pressure switch
and short-circuit the contacts. The switch state of DI06
should react on the display.
If it does, replace the pressure switch.
If it does not, use a multimeter to check that there are no
interruptions in the cable. Replace the I/O card if the cable is
OK.
Possible defect:
Pressure sensor,
cable or I/O card
Leakage
The leak detector
has been triggered.
Leaks in the
measuring device if
the springs of the
leak detector have
been bridged.
Possible defect:
Leak detector, cable,
I/O card
Malfunction
Peltier
The Peltier cooler
deviates > 3 °C from
the set point.
Following servicing
or maintenance, in
the event of high
ambient
temperatures, in the
event of unfavorable
suction conditions of
the ventilator
Possible defect:
Ventilator failure,
cable defective,
power supply outage
IR detector
malfunction
The measuring
signal of the IR
detector has failed.
f < 10000 Hz
After a brief power
outage.
Possible defect:
Cable, I/O card, IR
detector
Endress+Hauser
• Check for leaks.
• Check the leak alarm for an electrical short-circuit of the
spring contacts.
• Check signal processing (I/O card slot no. 29, switch input
DI05)
Go to the programming module PRG/TEST/INPUTS/SWITCH
INPUTS. Plug the cable for the carrier gas pressure switch into
slot 29. The display should change at switch input DI05 when
the connecting contacts on the pressure switch are bridged
manually.
If so, replace the leak detector.
If not, replace the I/O card.
Green LED on → Peltier cooler at operating temperature.
Red > °C LED on → Peltier cooler is too warm.
Red < °C LED on → Peltier cooler is too cold, control system is
defective.
• If no LED is lit, check the power supply of the Peltier cooler
control system.
• If the green LED is on, check the transmission cable to the I/O
card and the I/O card itself. Replace the I/O card if the cable is
OK.
• If the red LED > °C is on, check the function of the ventilator
on the cooler. Can the ventilator not draw in a sufficient
amount of air? Is the air temperature too high?
• Is the red LED < °C on?
Peltier cooler is too cold, control system is defective, replace
the Peltier cooler control system.
The IR detector switches to an automatic warm-up phase
following a power outage. The IR detector does not supply any
power during this time. The phase is finished after approx. 30 s
and the analyzer switches automatically to the measuring
mode.
• In the event of an error (malfunction constantly present after
60 s).
• Check the connecting cable between the I/O card and the IR
detector.
• Replace the cable of the gas card with another cable (e.g. pH
amplifier to the I/O card and check the signal input. Go to the
programming module PRG/TEST/INPUTS/ANALOG INPUTS
(I/O card slot No. 24 FI2 frequency input).
• Replace the IR detector if the I/O card is OK.
61
Diagnostics and troubleshooting
TOCII CA72TOC
Message
Cause
Tests or remedial action
Acid failure
If the pH value
permanently
deviates more than
±2.5 from the set
point.
Severely fluctuating
buffer capacity
values
• Check the acid cistern
• Is the acid concentration sufficient? Does the acid pump
operate at a maximum feed rate of 200 %?
Increase the acid concentration in the feeder.
• Is acid being dosed?
In the program module PRG/TEST/OUTPUTS/PUMPS, test
pump P3 by specifying values manually.
Check the pump hose for leaks.
• Adjust the pH sensor.
• Check signal processing (I/O card slot No. 26 frequency input
FI4): Disconnect the modular jack on the I/O card slot No. 26
- does the measured value drop?
Replace the I/O card if there is no change in the display value.
Possible defect:
Cable interruption,
pump hose, leaks,
pump control, pH
measurement
Unstable dosing
The drop monitor
does not count any,
or too few drop
events
• Is sample in the separation chamber?
• Is pump P2 pumping medium?
Can dripping be observed at the dosing head?
• Is the pressure sensor OK?
Go to the programming module PRG/TEST/INPUTS/ANALOG
INPUTS and observe the pressure sequence when medium
drips. Can a pressure increase > 10 mbar be observed?
Is the furnace fitted with the combustion pipe insert?
• Check signal processing: Remove the connector at MI4 (cable
53) and plug it back in. If the display has frozen, there is a
malfunction in the I/O card. Switch off the main switch, wait
a few seconds and switch it back on. If the error persists,
replace the I/O card.
Failure of water
pressure
The pressure sensor
for monitoring the
water supply has
been triggered.
Water pressure < 1
bar
• Check the water supply.
• Check signal processing (I/O card slot no. 35 switch input
DI03):
Go to the programming module PRG/TEST/INPUTS/SWITCH
INPUTS. Release the connecting cable on the pressure switch
and short-circuit the contacts. The switch state of DI03
should react on the display.
If it does, replace the pressure switch. If it does not, use a
multimeter to check that there are no interruptions in the
cable.
Replace the I/O card if the cable is OK.
Possible defect:
Pressure monitor,
cable, I/O card
Gas circuit
pressure high
The pressure sensor
measures a high
pressure in the gas
circuit.
A blockage is
forming in the gas
circuit.
Possible defect:
Pressure sensor,
cable, I/O card
Gas circuit
pressure too
high
The pressure sensor
measures too high a
pressure in the gas
circuit.
A blockage has
formed in the gas
circuit.
• Check the gas circuit for blockages. In particular, check the
acid filter, water trap, reactor and, if necessary, the heated
salt trap. Has the gas circuit flow rate fallen below 0.7 l/min?
Eliminate the blockage.
• Is the pressure sensor OK?
Go to the programming module PRG/TEST/INPUTS/ANALOG
INPUTS and observe the pressure sequence. Increase the
pressure by manually squeezing the gas circuit hose for
dosing. Can a pressure increase be observed?
• Check signal processing:
Is the plug correctly inserted in the multi in (MI) on the I/O
card?
Remove connection MI4 (cable 53) and plug it back in again.
If the display has frozen, there is a malfunction in the I/O
card. Switch the main switch off and on - if the error persists,
replace the I/O card.
Possible defect:
Pressure sensor,
cable, I/O card
62
Endress+Hauser
TOCII CA72TOC
Diagnostics and troubleshooting
Message
Cause
Tests or remedial action
Measured value
> measuring
range
The IR detector
continuously
returns a signal
that is greater than
the specification.
• The measured values in the flow of sample are constantly
higher than the measuring device configuration.
• When option pre-dilution is installed: Stop of dilution
function.
Faulty
adjustment
(adjustment
error 1)
CO2 concentrations
• Is the gas circuit leak-tight?
Check the gas-tightness of the analyzer.
• Replace the standard solutions.
Repeat the adjustment.
measured for standard C1 or C2 are
above the measuring range for the IR
detector.
Incorrect standard
solution
Possible defect:
Gas circuit leaking
Faulty
adjustment
(adjustment
error 2)
Calculated X0 value
is above the maximum value permitted for the IR detector used.
Gas circuit leaking,
standard solutions
Endress+Hauser
• Is the gas circuit leak-tight?
Check the gas-tightness of the analyzer.
• Check the adjustment values in the maintenance record.
Do one of the two record values deviate from the typical
value?
• Replace the standard solutions.
Faulty
adjustment
(adjustment
error 3)
The calibration line
slope is negative or
zero.
CO2 concentration
measured for standard 1 is higher
than that for standard 2
MV1, MV4, standard
solutions, container
empty
• Verify that the solenoid valves MV1 and MV4 are functioning
correctly.
In the program module PRG/TEST/OUTPUTS/BINARY
OUTPUTS, confirm the output SA1 for MV1 and SA4 for
MV4. If the solenoid valves do not switch, replace the
relevant solenoid valve.
• Have the standard solutions been prepared properly?
• Do the two standard solutions have the same concentration?
• Have the standard solutions been swapped?
• Are the standard canisters empty?
Faulty
adjustment
(adjustment
error 4)
KP value is smaller
than 30 or greater
than 150
Gas circuit leaking,
standard solutions
• Is the gas circuit leak-tight?
Check the gas-tightness of the analyzer.
• Have the standard solutions been prepared properly?
Replace the standard solutions.
• Biological growth in the standard canisters.
Replace the canisters.
• Dilution option - feed rate of pump P4 deviates from the
values determined. Use the program module SERVICE/
PUMPS/P1/P4 Replace pump hose to determine the feed rate
for pump P4.
Faulty
adjustment
(adjustment
error 5)
CO2 concentration <
min. permitted CO2
value.
(~ -9.4 % gas card
measuring range)
• Is the IR detector OK?
Convey pure supply gas through the IR detector.
Go to the programming module PRG/TEST/INPUTS/ANALOG
INPUTS and check whether the IR detector displays a negative
offset.
Replace the IR detector if the frequency displayed has
dropped below 10000 Hz.
63
Diagnostics and troubleshooting
TOCII CA72TOC
Message
Cause
Tests or remedial action
Fault CO2 base-
Baseline value 5 %
of IR detector full
scale value
New catalyst, pellets
for soda-lime
scrubbers are used
up, gas generator
malfunction,
defective carrier gas
valve, leak in gas
circuit, pump
adjustment for pump
P2 is no longer upto-date
• Once the catalyst has been replaced, the catalyst may outgas.
This can result in an error message, particularly in low CO2measuring ranges. The fault resolves itself after a few
measuring cycles.
• Are the pellets completely discolored?
Replace the pellet filling.
• Verify that the gas generator is functioning correctly.
• Does the carrier gas valve allow sufficient gas through to
allow for adequate flushing?
• Is the carrier gas valve sealed?
• Carry out a pump adjustment for pump P2.
• Is the gas circuit leak-tight?
Carry out a leakage test.
Input error
C1>C2
Input value for C1 is
greater than for C2
• Enter the correct concentrations.
Calibration
marked with an
asterisk
IR signal is less than
75 % of the value for
C2 from the last
adjustment
• Replace the standard solution for C2. Repeat the calibration.
Internal comfault 1
I/O not responding
during INIT process
• Switch off the main switch and, after a short period of time,
switch it back on again. If the error persists, contact E+H
Service.
Internal comfault 2
I/O not responding
during NOINIT
process
• Switch off the main switch and, after a short period of time,
switch it back on again. If the error persists, contact E+H
Service.
Internal comfault 10
KEYBOARD not
responding
• Switch off the main switch and, after a short period of time,
switch it back on again. If the error persists, contact E+H
Service.
Internal comfault 20
CRC error between I/
O – CPU or between
KEYBOARD – CPU
• Switch off the main switch and, after a short period of time,
switch it back on again. If the error persists, contact E+H
Service.
line
64
Endress+Hauser
TOCII CA72TOC
Maintenance
8
Maintenance
To ensure efficient operation of the analyzer, maintenance must be carried out at regular
intervals.
Before attempting to service the device, you must be completely familiar with the
maintenance processes and have perfectly understood the tasks involved. All the
maintenance tasks described in this chapter should only be carried out by suitably
trained specialists. Mistakes made during maintenance can result in inaccurate
operation and pose a safety hazard.
Interval
Maintenance task
At least once per week
• Visual inspection
• Check sample conditioning (see corresponding Operating Instructions)
At least once per month
• Check feed rate of pump P1/P4 and P2
• Replace standard
Every 3 months at the latest
•
•
•
•
•
If salt quantity > 1 g/l, then
perform the following every 3
months at the latest:
• Replacing the acid filter
• Replace catalyst
• Clean combustion pipe
1 x per year
• Replace filter pads in ventilators (do not clean)
• Replace membrane filter (gas filter)
Clean strip and separation chamber
Replace glass bead
Adjust pH sensor
Replace pump hoses
Check the filter pads in the ventilators and replace if necessary.
The maintenance intervals depend greatly on your application. For this reason, adapt the
maintenance intervals to your specific requirements but make sure to perform these
maintenance tasks regularly!
Endress+Hauser
65
Maintenance
TOCII CA72TOC
Flow chart
Use this chart as reference if you have to replace or reconnect pipes or tubes.
a0012369-en
Fig. 25:
66
Flow chart
Endress+Hauser
TOCII CA72TOC
Maintenance
Interior views
a0011861
Fig. 26:
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
!
Display and operating elements
Main switch
Display and operating elements
USB connection
Pump P2 sample analysis
Junction box
Compressor switch
Ventilator
Separation chamber
Dosing valve
Ventilator
Injection unit
Acid filter
Tube furnace with catalyst
Air outlet (filter pad)
Combined filter (water trap)
16
17
18
19
20
21
22
23
24
25
26
27
28
29
Heated salt trap (optional)
Circuit gas flowmeter
Dilution water pump P5 (optional)
Acid hose connector
MV4 calibration standard changeover C1/C2
MV1 wastewater/calibration standard
changeover
Valve for online sample/manual sample
Pump P3, acid dosing
Pump P4, sample - dilution (optional)
Mixing chamber (optional)
Pump P1, sample - strip chamber/condensate
extraction
Condensate hose connector
Strip chamber with pH electrode
Sample conditioning
WARNING
The device is energized.
Touching current-carrying components may result in injury or death.
‣ First switch off the main switch. Then open the rear of the device with the special key
provided.
Endress+Hauser
67
Maintenance
TOCII CA72TOC
a0012022
Fig. 27:
1
2
3
4
5
6
7
8
9
10
11
12
13
68
Rear of device
Peltier cooler
Peltier cooler - glass insert
Motor controller P1 to P6
Peltier cooler - control unit
Membrane compressor
Motor P2 sample - analysis
Computer unit
Power distribution
Magnetic stirrer control
pH - amplifier
MV7 carrier gas flushing
MV3 stripping gas supply
Gas connection block
14
15
16
17
18
19
20
21
22
23
24
Restrictor stripping gas
Restrictor carrier gas
Motor P1, sample - strip chamber /
condensate drainage
Check valve
Motor P4, sample - dilution (optional)
Motor P3 acid dosing
MV5 bypass flushing
Water connection manifold
IR detector
Leak detector contacts
Motor P5, dilution water pump (optional)
Endress+Hauser
TOCII CA72TOC
Maintenance
8.1
Cleaning the housing
Clean the analyzer housing as follows:
• Use a fluoride-free and solvent-free cleaning agent and a lint-free cloth.
8.2
!
Visual inspection
WARNING
There is a risk of infection when working with wastewater!
‣ For this reason, wear protective goggles, protective gloves and protective clothing.
!
CAUTION
There is a risk of injury due to hot components!
‣ Wear heat-resistant gloves when in contact with hot components in the vicinity of the
combustion furnace..
Carry out the visual inspection at least once a week:
• Check whether the measurement is operating in the standard range.
• Is the sample supply OK? To check, hold a collecting vessel under the valve and set the
valve briefly to "Manual sample". Does the sample flow out of the bypass?
• Is sample being dosed into the furnace?
• Are hoses P1 to P3 (optionally P4) leak-tight?
• Is sufficient C1 and C2 standard still available?
• Is sufficient stripping acid still available?
• If the condensate is being collected in a container, check if the container if full.
• Check the gas supply.
Pressure regulator at 2 bar (29 psi)?
Circuit gas (right flowmeter) at 0.7 to 1.2 l/min (0.18 to 0.32 gal/min)?
• Is the water supply pressure between 3 ±0.2 bar (43 ±3 psi)?
• Is the acid filter free of condensate and not severely discolored?
• Check stripping gas bubbling.
• Does the rotating slit filter rotate uniformly?
• Is a slit visible between the rotating body and the base of the chamber?
a0012724
Fig. 28:
Endress+Hauser
Rotating slit filter
a0012723
Fig. 29:
Rotating slit filter, detailed view
69
Maintenance
TOCII CA72TOC
8.3
!
Topping up reagents
CAUTION
There is a risk of injury when working with reagents.
‣ Observe the warnings on the safety data sheets.
‣ Wear acid-resistant protective gloves, protective goggles and a protective laboratory
coat!
‣ In addition, adhere to the regulations applicable in your country and to your line of work
governing the handling of caustic chemicals. Rinse off splashes immediately with plenty
of water and 1 % sodium hydrocarbonate solution. Consult a physician and show him/
her the information on the canister.
Only use original canisters for measurement. No liability is accepted for damage resulting
from the use of other chemicals.
The stripping reagent and standards C1 and C2 must be topped up or replaced. To prepare
the reagents, see the "Preparing the chemicals" section.
Parent solution
• 5 000 mg/l TOC; for preparing standard solutions 1 and 2
Order number:
1000 ml (33.8 fl.oz.): CAY451-V10C01AAE
• 100 000 mg/l TOC; for preparing standard solutions 1 and 2
Order number:
1000 ml (33.8 fl.oz.): CAY451-V10C10AAE
Stripping reagent
• Stripping reagent for CA72TOC
Order number:
1000 ml (33.8 fl.oz.): CAY450-V10AAE
70
Endress+Hauser
TOCII CA72TOC
Maintenance
8.4
Replacing the standard
This section describes how the standard solutions are replaced.
Step
Display
1
2
Description
BYPASS SCREEN stops the measuring
operation.
BYPASS-SCREEN
Change the standard solution as
follows:
1. Remove the canister for the
standard solution.
MAINTENANCE COMPLETED ?
2. Dispose of any remaining
standard solution.
3. Rinse the canister with deionized
water.
4. Fill up the canister with new
standard solution.
[E] TO CONFIRM
5. Position the canister at the
designated place.
6. Press the [E] key.
Operation starts.
If you have changed the concentration of the standard solution, enter the new concentration
in the "Measuring range data" menu item and then trigger an adjustment.
Endress+Hauser
71
Maintenance
TOCII CA72TOC
8.5
Entering the concentration of the standard solution
This section describes how the concentration of the standard solution is entered.
Select the programming mode for this purpose:
Step
Display
Description
1
2
72
• Enter the values for STANDARD C1
and STANDARD C2.
Endress+Hauser
TOCII CA72TOC
Maintenance
8.6
Service menu
The maintenance tasks are supported by the service software. This menu software is split
into four sections:
Endress+Hauser
Pumps
Replace hose P1/P4
Replace hose P2
Replace hose P3
Adjust P2
Calibration
Analyzer adjustment
Analyzer calibration
Empty volume dosing
Adjustment pH sensor
Cleaning
Screen flush
Power flush
Bypass screen
Strip and separation chamber
Opening the circuit
Combustion pipe
Leakage test
Filter
Replacing the acid filter
Replacing the gas filter
Replacing the O2 filter
Replacing the heated filter
73
Maintenance
TOCII CA72TOC
8.7
Pumps
8.7.1
Replacing the hoses on pumps P1 and P4
This section describes how the hoses of pumps P1 and P4 are replaced and the delivery rate
is determined.
!
CAUTION
Danger of crushing!
‣ Never reach into the pump head while the pump is in operation!
P1
P4
a0012483
Fig. 30:
Position of pumps P1 and P4
You require the following parts:
• 10 ml graduated cylinder
• 2.5 mm Allen key
• Dosing nozzle (injector)
• Paper to absorb liquids
• Collecting vessel with volume of approx. 150 ml (5 fl.oz) to collect liquid
• Silicone grease
74
Endress+Hauser
TOCII CA72TOC
Maintenance
4
5
3
2
6
7
1
8
a0012364
Fig. 31:
1
2
3
4
Pump P1
Pump retainer
Pump head
Hose cassette (sample hose)
Hose cassette (condensate hose)
5
6
7
8
Motor
Condensate hose (BK-BK) 0.76 mm (0.03") ID
Sample hose (VT-WH) 2.79 mm (0.11") ID
Adjusting screw
Pump P1: The sample hose to the strip chamber has an internal diameter of 2.79 mm (0.11")
with violet-white (VT-WH) color coding and the condensate drain hose has an ID of 0.76
mm (0.03") with black-black (BK-BK) color coding.
Pump P4 (only for "With predilution" version): sample hose to the static mixer 2.79 mm
(0.11") with VT-WH color coding.
Step
Display
Description
1
Endress+Hauser
75
Maintenance
TOCII CA72TOC
Step
Display
2
Description
!
WARNING
There is a risk of infection when
working with wastewater!
Wear protective gloves, goggles
and protective clothing.
‣
Press the [E] key.
The strip and separation chambers
are flushed with pressurized water.
3
REPLACE HOSE P1/4
SWITCH VALVE TO MANUAL SAMPLING,
PUT COLLECTING VESSEL BELOW
1. Set the sample valve to manual
sampling.
2. To empty the pump hose and the
strip chamber, place a collecting
vessel under the hose connection
for manual sampling.
3. Press the [E] key.
[E] TO CONFIRM
4
For versions "with predilution":
1. First open the hose cassettes for
pump 1.
2. Now open the hose cassette for
pump 4.
3. Drain the strip chamber.
4. Press the [E] key.
76
Endress+Hauser
TOCII CA72TOC
Maintenance
Step
Display
Description
For versions "without predilution":
1. Open the hose cassette for pump
1.
2. Drain the strip chamber.
3. Press the [E] key.
5
REPLACE HOSE P1/4
In areas with a low amount of dust:
Grease the new hoses with a light
coating of silicone grease.
1. Hold paper under the strip
chamber and mixer hose
connection.
REPLACE PUMP HOSE
2. Release the pump hoses (items 6
+ 7) from the connections.
3. Replace the hoses.
4. Press the [E] key.
[E] TO CONFIRM
6
For versions "with predilution":
REPLACE HOSE P1/4
CLOSE HOSE CASSETTE(S) P1
CLOSE HOSE CASSETTE P4
CONNECT LOWER HOSES
FILL HOSE WITH SAMPLE
[>] START/STOP PUMP 4 (100%)
[E] TO CONFIRM
Continue with Step 6 for versions
"with predilution".
Continue with Step 14 for versions
"without predilution".
1. Lock the hose cassettes in place in
the retainer.
2. Make sure the hose cassette is
seated correctly in the retainer.
3. Connect the P1 and P4 suction
sides.
4. Press the [>] key to fill the hose
with sample.
5. Check the drip pattern (even feed
rate).
You can now check the contact
pressure of the hose cassette:
1. Release the adjusting screw
(item 8) until no more medium is
conveyed.
2. Now tighten the adjusting screw
until the unit starts to pump
medium again. Even medium
delivery should be visible across
all pump heads.
3. Tighten the adjusting screw one
revolution.
Endress+Hauser
77
Maintenance
TOCII CA72TOC
Step
7
Display
REPLACE HOSE P1/4
Description
If required, you can measure the
capacity of pump hose P4.
1. Secure or hold the 10 ml
graduated cylinder close to pump
P4.
IF REQUIRED, MEASURE CAPACITY OF HOSE P4
PREVIOUS VOLUME DELIVERED x.x ml
2. Guide hose P4 into the graduated
cylinder.
3. Press the [>] key so pump P4 is
operating at 100% capacity for 60
s.
[>] START/STOP PUMP 4 (60sec)
[E] TO CONFIRM
You can enter the new value after 60
seconds.
The value typically lies between 5.5
and 7 ml (0.18 and 0.24 fl.oz).
REPLACE HOSE P1/4
The amount entered is displayed so
you can check it is correct.
Press the [E] key.
NEW VOLUME DELIVERED x.x ml
[>] START/STOP PUMP 4 (60sec)
[E] TO CONFIRM
8
REPLACE HOSE P1/4
SHUT STRIP CHAMBER INLET WITH
SEPARATE PLUG
1. Seal the strip chamber inlet with
a separate plug (e.g. plug for leak
test).
2. Connect hose P4 at the pressure
side.
3. Press the [E] key.
CONNECT HOSE P4
[E] TO CONFIRM
78
Endress+Hauser
TOCII CA72TOC
Maintenance
Step
9
Display
Description
1. Use the dosing nozzle of the
injector to expand the hose end of
the condensate hose.
2. Connect the suction side of the
condensate hose and place the
pressure side of the condensate
hose into a glass containing
water.
3. Press the [>] key to fill the
sample hose with sample.
4. Check the drip pattern of the
sample hose and check the air
bubbles in the glass of water
(even feed rate).
You can now check the contact
pressure of the hose cassette:
1. Release the adjusting screw
(item 8) until no more medium is
conveyed (air bubbles).
2. Now tighten the adjusting screw
until the unit starts to pump
medium again. Even medium
delivery should be visible across
all pump heads.
3. Tighten the adjusting screw one
revolution.
4. Reconnect the pressure side of
the condensate hose.
10
If required, you can measure the
capacity of pump hose P1.
1. Secure or hold the 10 ml
graduated cylinder close to pump
P1.
2. Guide hose P1 into the graduated
cylinder.
3. Press the [>] key so pump P1 is
operating at 100% capacity for 60
s.
You can enter the new value after 60
seconds.
The value typically lies between 5.5
and 7 ml (0.18 and 0.24 fl.oz).
Endress+Hauser
79
Maintenance
TOCII CA72TOC
Step
Display
Description
Press the [E] key.
11
1. Connect hose P1 at the pressure
side.
2. Press the [E] key.
12
1. Set the manual valve on the
bypass to sampling.
REPLACE HOSE P1/4
SWITCH VALVE TO BYPASS
2. Pump the sample out of the
bypass with pump P4 until
dilution.
3. Press the [E] key.
FILL HOSE WITH SAMPLE
[>] START/STOP PUMP 4 (100%)
[E] TO CONFIRM
13
Stabilize dilution
REPLACE HOSE P1/4
Continue with Step 18.
XXX
STABILIZE DILUTION
80
Endress+Hauser
TOCII CA72TOC
Maintenance
Step
14
Display
Description
For versions "without predilution":
1. Lock the hose cassettes in place in
the retainer.
2. Make sure the hose cassette is
seated correctly in the retainer.
3. Use the dosing nozzle of the
injector to expand the hose end of
the condensate hose.
4. Connect the P1 suction side.
5. Set the sample tap to bypass.
6. Place the pump side of the
condensate hose into a glass of
water.
7. Press the [>] key to fill the hose
with sample.
8. Check the drip pattern of the
sample hose and check the air
bubbles in the glass of water
(even feed rate).
You can now check the contact
pressure of the hose cassette:
1. Release the adjusting screw
(item 8) until no more medium is
conveyed (air bubbles).
2. Now tighten the adjusting screw
until the unit starts to pump
medium again. Even medium
delivery should be visible across
all pump heads.
3. Tighten the adjusting screw one
revolution.
4. Reconnect the pressure side of
the condensate hose.
15
If required, you can measure the
capacity of pump hose P1.
1. Secure or hold the 10 ml
graduated cylinder close to pump
P1.
2. Guide hose P1 into the graduated
cylinder.
3. Press the [>] key so pump P1 is
operating at 100% capacity for 60
s.
You can enter the new value after 60
seconds.
The value typically lies between 5.5
and 7 ml (0.18 and 0.24 fl.oz).
Endress+Hauser
81
Maintenance
TOCII CA72TOC
Step
Display
Description
Enter the new feed rate.
Press the [E] key.
16
1. Connect hose P1 at the pressure
side.
2. Press the [E] key.
17
1. Set the manual valve to bypass.
2. Using pump P1, pump the sample
as far as the strip chamber.
3. Press the [E] key.
18
82
Automatic filling of the strip
chamber, conditioning of the strip
chamber with active acid dosing
Endress+Hauser
TOCII CA72TOC
Maintenance
8.7.2
Replacing the hose on pump P2
This section describes how the hose of pump P2 is replaced. The following service menus are
automatically started when this service menu finishes:
1.
Adjusting pump P2
2.
Empty volume dosing
!
CAUTION
Danger of crushing!
‣ Never reach into the pump head while the pump is in operation!
a0012484
Fig. 32:
Position of pump P2
You require the following parts:
• 10 ml graduated cylinder
• 2.5 mm Allen key
• Dosing nozzle (injector)
• Paper to absorb liquids
• Collecting vessel with volume of approx. 150 ml (5 fl.oz) to collect liquid
• Silicone grease
Endress+Hauser
83
Maintenance
TOCII CA72TOC
6
5
4
2
1
3
a0014230
1
2
3
Pump retainer
Pump head
Sample hose (BK-BK) 0.76 mm (0.03") ID
4
5
6
Adjusting screw
Hose cassette
Motor
Pump P2: Pumps sample from the separation chamber into the furnace.
Hose: (BK-BK) 0.76 mm (0.03")
Step
Display
Description
1
2
!
WARNING
When working with wastewater,
there is a risk of infection!
Wear protective gloves, goggles
and protective clothing.
‣
Press the [E] key.
84
Endress+Hauser
TOCII CA72TOC
Maintenance
Step
3
Display
Description
The hose is drained automatically.
REPLACE HOSE P2
XX
PLEASE WAIT
PUMP CONVEYS BACKWARDS
[E] TO CONFIRM
4
REPLACE HOSE P2
1. Open the cover of the separation
chamber.
2. Empty the separation chamber
with the syringe (part of the
scope of supply).
EMPTY SEPARATION CHAMBER USING NOZZLE!
3. Press the [E] key.
[E] TO CONFIRM
5
REPLACE HOSE P2
REPLACE PUMP HOSE
In areas with a low amount of dust:
Grease the new hoses with a light
coating of silicone grease.
1. Release the hose at the injection
unit and at the separation
chamber.
2. Disengage the hose cassette.
3. Change the hose.
4. Use the dosing nozzle of the
injector to expand the hose end.
5. Engage the hose cassette.
[E] TO CONFIRM
6
6. Press the [E] key.
1. Seal the separation chamber.
2. Connect the hose at the suction
side.
3. Start the service menu "Adjust
pump P2" or press the [E] key to
quit the service menu.
Endress+Hauser
85
Maintenance
TOCII CA72TOC
Step
7
Display
REPLACE HOSE P2
Description
1. Press the [>] key to fill the hose
with sample.
2. Check the drip pattern (even feed
rate).
FILL HOSE WITH SAMPLE
If a steady flow of medium does not
occur, check the contact pressure of
the hose cassette:
1. Release the adjusting screw
(item 4) until no more medium is
conveyed.
[>] START/STOP PUMP 2 (100%)
[E] TO CONFIRM
2. Now tighten the adjusting screw
until the unit starts to pump
medium again. Even medium
delivery should be visible across
all pump heads.
3. Tighten the adjusting screw one
revolution.
Press the [E] key.
8
1. Connect the hose P2 to the
injection unit
2. Press the [E] key.
Measuring operation is started
automatically.
The accuracy of the feed rate of pump P2 affects the measurement result. For this
reason, start the following service menus in the order indicated once the pump hose has
been replaced:
1."Adjust pump 2"
2."Empty volume dosing"
Please also note that the hoses are subject to deterioration in the first hours of
operation. For this reason, repeat these maintenance tasks after 24 hours.
86
Endress+Hauser
TOCII CA72TOC
Maintenance
8.7.3
Replacing the hose on pump P3
This section describes how the hose of pump P3 is replaced.
!
CAUTION
Danger of crushing!
‣ Never reach into the pump head while the pump is in operation!
!
CAUTION
Risk of injury due to acid
‣ Wear acid-resistant protective gloves, protective goggles and protective clothing!
NOTICE
Failure to comply may result in incorrect measurements
‣ No medium containing TOC should enter the acid feeder!
a0012485
Fig. 33:
Position of pump P3
You require the following parts:
• Acid-resistant protective gloves, protective goggles and protective clothing
• 10 ml graduated cylinder
• 2.5 mm Allen key
• Dosing nozzle (injector)
• Paper to absorb liquids
• Collecting vessel with volume of approx. 150 ml (5 fl.oz) to collect liquid
• Silicone grease
Endress+Hauser
87
Maintenance
TOCII CA72TOC
3
4
2
5
1
6
a0012368
Fig. 34:
1
2
3
Pump P3
Pump retainer
Pump head
Hose cassette
4
5
6
Motor
Acid hose (BK-BK) 0.76 mm (0.03") ID
Adjusting screw
Pump P3: Pumps acid from the feeder tank into the strip chamber.
Hose: (BK-BK) 0.76 mm (0.03")
Step
Display
Description
1
2
!
CAUTION
Risk of injury due to acid!
Wear acid-resistant protective
gloves, protective goggles and
protective clothing. Rinse areas
splashed with acid immediately
with plenty of water and 1 %
sodium hydrocarbonate solution.
Consult a physician and show
him/her the information on the
canister.
‣
Press the [E] key.
The strip and separation chambers
are flushed with pressurized water.
88
Endress+Hauser
TOCII CA72TOC
Maintenance
Step
3
Display
REPLACE HOSE P3
EMPTY STRIP CHAMBER BY
DISCONNECTING HOSE CONNECTION OF
SAMPLE PUMP P1 AT THE STRIP CHAMBER
PLACE VESSEL BELOW TO COLLECT
CONTENT OF STRIP CHAMBER
Description
1. Release the hose connection from
pump P1 to the strip chamber.
2. Have a vessel at the ready to
collect the liquid.
3. Drain the strip chamber.
4. Press the [E] key.
[E] TO CONFIRM
4
REPLACE HOSE P3
DANGER! ACID
REMOVE ACID SUCTION HOSE FROM
ACID FEEDER AND PLACE VESSEL
BELOW TO COLLECT ACID
RELEASE HOSE CASSETTE P3 AND EMPTY
ACID LINE
Caution! Acid!
1. Remove the acid suction hose
from the acid cistern and place
the end in a collecting vessel.
2. Release the hose cassette.
3. Make sure that the acid runs back
into the collecting vessel.
NOTICE
Failure to comply may result in
incorrect measurements.
No medium containing TOC
should enter the acid feeder!
‣
[E] TO CONFIRM
5
4. Press the [E] key.
In areas with a low amount of dust:
Grease the new hoses with a light
coating of silicone grease.
1. Release the pump hose from the
connections.
2. Change the hose.
3. Press the [E] key.
6
REPLACE HOSE P3
1. Connect the hose of pump P1 to
the strip chamber.
2. Press the [E] key.
CONNECT SAMPLE HOSE PUMP P1
TO STRIP CHAMBER
[E] TO CONFIRM
Endress+Hauser
89
Maintenance
TOCII CA72TOC
Step
Display
Description
7
1. Rinse the suction hose of the acid
pump.
2. Guide the suction tube into the
acid feeder tank.
3. Insert the hose cassette into P3
and connect the hose to the hose
connector.
4. Use the dosing nozzle of the
injector to expand the hose end.
5. Press the [>] key until acid enters
the strip chamber.
6. Make sure medium is being
conveyed evenly in the suction
hose.
You can now check the contact
pressure of the hose cassette:
1. Release the adjusting screw
(item 6) until no more medium is
conveyed.
2. Now tighten the adjusting screw
until the unit starts to pump
medium again. Even medium
delivery should be visible across
all pump heads.
3. Tighten the adjusting screw one
revolution.
Continue with Step 8 for versions
"with predilution".
Continue with Step 10 for versions
"without predilution".
8
Dilution stabilizes for 120 s.
REPLACE HOSE P3
XXX
STABILIZE DILUTION
9
REPLACE HOSE P3
XX
FILL HOSE WITH SAMPLE
90
Endress+Hauser
TOCII CA72TOC
Maintenance
Step
10
Display
Description
Automatic filling of the strip
chamber, conditioning of the strip
chamber with active acid dosing
REPLACE HOSE P3
XX
Measuring operation is started
automatically.
FILL STRIP AND SEPARATION CHAMBERS
Endress+Hauser
91
Maintenance
TOCII CA72TOC
8.7.4
Adjusting pump P2
This section describes how the feed rate of pump P2 is determined. The "Empty volume
dosing" service menu is started automatically when this service menu finishes:
!
CAUTION
Danger of crushing!
‣ Never reach into the pump head while the pump is in operation!
a0012484
Fig. 35:
Position of pump P2
You require the following parts:
• 10 ml graduated cylinder
• 2.5 mm Allen key
• Dosing nozzle (injector)
• Paper to absorb liquids
• Collecting vessel with volume of approx. 150 ml (5 fl.oz) to collect liquid
92
Endress+Hauser
TOCII CA72TOC
Maintenance
6
5
4
2
1
3
a0014230
Fig. 36:
1
2
3
Pump P2
Pump retainer
Pump head
Sample hose (BK-BK) 0.76 mm (0.03") ID
Step
Display
4
5
6
Adjusting screw
Hose cassette
Motor
Description
1
2
!
CAUTION
There is a risk of infection when
working with wastewater!
For this reason, wear protective
goggles, protective gloves and
protective clothing.
‣
Press the [E] key.
Endress+Hauser
93
Maintenance
TOCII CA72TOC
Step
3
Display
Description
1. Release hose P2 at the dosing
needle.
2. Hold a collecting vessel under the
hose.
3. Start pump P2 with the [>] key.
4. Fill the hose until an even
amount of sample exits the hose.
No air bubbles should be
conveyed. Make sure that even
dosage takes place over all the
rollers of the pump head.
5. Stop the pump as soon as a steady
flow of medium is pumped.
If a steady flow of medium does not
occur, you can check the contact
pressure of the hose cassette:
1. Release the adjusting screw
(item 4) until no more medium is
conveyed.
2. Now tighten the adjusting screw
until the unit starts to pump
medium again. Even medium
delivery should be visible across
all pump heads.
3. Tighten the adjusting screw one
revolution.
4
1. Place a graduated beaker
underneath the end of the dosing
tube.
2. Press the [E] key to start timecontrolled dosage.
5
94
Pump P2 pumps at 100 % for 10
minutes.
Endress+Hauser
TOCII CA72TOC
Maintenance
Step
6
Display
Description
ADJUSTMENT PUMP 2
ENTER ml ?
Enter the delivery volume
determined.
The value typically lies between 8.5
and 9.5 ml (0.29 and 0.32 fl.oz).
[E] TO CONFIRM
7
Press the [E] key.
8
1. Connect hose P2.
2. Press the [E] key (the "Empty
volume dosing" service menu is
started).
The accuracy of the feed rate of pump P2 affects the measurement result25
For this reason, the "Empty volume dosing" service menu is started once pump P2 has
been adjusted.
Endress+Hauser
95
Maintenance
TOCII CA72TOC
8.7.5
Cleaning dilution water pump P5
If you use deionized water as the dilution medium, pump P5 only has to be cleaned as
part of the annual maintenance work performed by Endress+Hauser Service.
If you are using drinking water as the dilution medium, the maintenance intervals may
be reduced depending on the water hardness level. In this case, contact Endress+Hauser
Service
a0014102
Fig. 37:
96
Position of pump P5
Endress+Hauser
TOCII CA72TOC
Maintenance
8.8
Adjusting
8.8.1
Adjusting the analyzer
This section describes how the analyzer is adjusted.
The analyzer is adjusted using two-point adjustment. Two different standard solutions are
measured for this purpose. These standard solutions are connected in the analyzer.
Adjustment procedure:
1.
The basic line is measured.
2.
The analyzer measures the concentration of standard C1.
3.
The basic line is measured.
4.
The analyzer measures the concentration of standard C2.
5.
The offset X0 and the slope Kp are calculated from these values.
Measuring signal =
CO2(C) = CO2(dose) - CO2(base line)
a0012547-en
Fig. 38:
Calculation of the measuring signal
X0 =
CO2(C1) - C1
CO2(C2) -
CO2(C1)
(C2 - C1)
a0012549
Calculation of the offset X0
Kp =
C2
CO2(C2) - X0
actual dosing volume
0.3125
a0012548-en
Fig. 39:
Calculation of the standardized slope Kp
Measured signal
Kp
X0
C1
C2
Concentration
a0012550-en
Fig. 40:
Adjustment curve
CO2(C1) = CO2 concentration of the measurement of standard C1.
CO2(C2) = CO2 concentration of the measurement of standard C2.
The offset and the reciprocal standardized slope of the adjustment curve (measuring signal
per concentration) are saved in the maintenance record log under ADJUSTMENT
CONSTANTS.
Analyzer adjustment can be started in three ways:
• Activated manually
• Activated remotely
• Activated automatically
Endress+Hauser
97
Maintenance
TOCII CA72TOC
Manually activated adjustment
Step
Display
Description
1
Remotely activated adjustment
Adjustment can be triggered by a floating contact. Use input 2 of the "binary in" terminal strip
for this purpose.
I
1
1
II
2
1
2
0/4-20
mA
+
0/4-20
mA
+
binary in
2
III
1
IV
binary out
2
40
-
41
-
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
3
4
5
6
7
8
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
RS 232
a0011911
Fig. 41:
1
2
3
4
98
Signal connection
Calibration, external activation
Adjustment, external activation
Screen flush, external activation
Power flush, external activation
5
6
7
8
Not assigned
Not assigned
Standby, external activation
Channel switchover, external activation
(optional)
Endress+Hauser
TOCII CA72TOC
Maintenance
Automatically activated adjustment
The analyzer can be adjusted automatically.
Select the programming mode for this purpose:
Step
Display
Description
1
2
Endress+Hauser
• For CAL./ADJUST. n DAYS, specify
the number of days after which an
adjustment is to be performed. It is
generally not necessary to select
more than one adjustment every
three days.
• Enter the value 2 for CAL./
ADJUSTM.
Value 1 = Calibration
Value 2 = Adjustment
99
Maintenance
TOCII CA72TOC
8.8.2
Calibrating the analyzer
This section describes how the analyzer is calibrated.
The analyzer checks the current recovery by measuring the standard solution C2. In contrast
to adjustment, the adjustment constants are not changed.
The standard solution C2 is connected to the analyzer.
Analyzer calibration can be started in three ways:
• Activated manually
• Activated remotely
• Activated automatically
Manually activated calibration
Step
Display
Description
1
Remotely activated calibration
Calibration can be triggered by a floating contact. Use input 1 of the "binary in" terminal strip
for this purpose.
I
1
1
II
2
1
2
0/4-20
mA
+
0/4-20
mA
+
binary in
2
III
1
IV
binary out
2
40
-
41
-
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
3
4
5
6
7
8
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
RS 232
a0011911
Fig. 42:
1
2
3
4
100
Signal connection
Calibration, external activation
Adjustment, external activation
Screen flush, external activation
Power flush, external activation
5
6
7
8
Not assigned
Not assigned
Standby, external activation
Channel switchover, external activation
(optional)
Endress+Hauser
TOCII CA72TOC
Maintenance
Automatically activated calibration
The analyzer can be calibrated automatically.
Select the programming mode for this purpose:
Step
Display
Description
1
2
• For CAL./ADJUST. n DAYS, specify
the number of days after which a
calibration is to be performed. It is
generally not necessary to select
more than one calibration every
three days.
• Enter the value 1 for CAL./
ADJUSTM.
Value 1 = Calibration
Value 2 = Adjustment
Analog value output during calibration
During calibration, the last TOC value is transmitted to the analog output and held until
the calibration value has been determined. The calibration value is then transmitted to
the analog output (or to analog output channel 1 in the two-channel version) until a
new measured value is determined for the current sample. During calibration, relay IV
(OPERATION CHECK) is opened until a new measured value is available in measuring
mode. If the analog output is used for control purposes, this OPERATION CHECK signal
can be used to declare the analog output as invalid.
Endress+Hauser
101
Maintenance
TOCII CA72TOC
8.8.3
Empty volume dosing
Use this service menu to determine the dead volume of pump P2 from the separation
chamber to the end of the capillary.
a0012487
Fig. 43:
Step
Position
Display
Description
1
2
The hose of pump P2 is drained.
3
The pump pumps automatically in the
direction of the capillary.
Pumping stops if:
a) A drop is detected or
b) The system times out after 180 s.
In the event of a), continue with Step
4.
In the event of b), continue with Step
6.
102
Endress+Hauser
TOCII CA72TOC
Maintenance
Step
4
5
6
Display
Description
The new volume value determined
appears on the display. The value,
including the volume and time, is
entered in the EMPTY VOLUME
DOSING log.
The value determined is saved in the
EMPTY VOLUME P2 [μl] parameter.
Measuring operation is started once you confirm with the [E] key.
"Empty volume dosing" ends.
The system has timed out.
Determine the empty volume
manually.
The service is restarted and the
automatic determination function is
disabled.
Empty volume dosing can now be
carried out manually.
Endress+Hauser
7
The pump is drained.
8
Start the pump with the [E] key.
103
Maintenance
TOCII CA72TOC
Step
Description
9
Stop the pump with the [E] key when
the first drop falls.
10
The new volume value determined
appears on the display. The value,
including the volume and time, is
entered in the EMPTY VOLUME
DOSING log.
The value determined is saved in the
EMPTY VOLUME P2 [μl] parameter.
11
104
Display
Measuring operation is started once you confirm with the [E] key.
"Empty volume dosing" ends.
Endress+Hauser
TOCII CA72TOC
Maintenance
8.8.4
Adjustment pH sensor
This section describes how to remove and install the pH sensor, and how to perform twopoint adjustment on the pH sensor.
a0012478
Fig. 44:
Position of the sensor
To adjust the pH sensor, you require:
• Deionized water
• Buffer solution pH = 4.00
• Buffer solution pH = 7.00
• Paper towels to absorb liquid
• Vessel to collect liquids
1
2
3
4
a0012176
Fig. 45:
1
2
3
4
Endress+Hauser
Strip chamber and separation chamber
pH sensor
Cover
Thread adapter nut
Strip chamber
105
Maintenance
TOCII CA72TOC
Step
Display
Description
1
2
1. Release the thread adapter nut
(3) on the strip chamber (4).
ADJUSTMENT pH SENSOR
2. Remove the cover (2) with the pH
sensor (1) from the strip
chamber.
RELEASE COVER OF STRIP CHAMBER
3. Press the [E] key.
CAREFULLY REMOVE SENSOR
[E] TO CONFIRM
3
1. Rinse the sensor with deionized
water.
2. Hold the sensor in the buffer
solution pH = 4.00.
3. Press the [E] key.
4
1. Wait until the measured value
stabilizes (a bar appears to the
right of the measured value).
ADJUSTMENT pH SENSOR
2. Press the [E] key.
MEASURED VALUE
X.XX
[E] TO CONFIRM
106
Endress+Hauser
TOCII CA72TOC
Maintenance
Step
Display
Description
5
1. Rinse the sensor with deionized
water.
2. Hold the sensor in the buffer
solution pH = 7.00.
3. Press the [E] key.
6
1. Wait until the measured value
stabilizes (a bar appears to the
right of the measured value).
ADJUSTMENT pH SENSOR
2. Press the [E] key.
Calculation of the calibration values.
Typical slope value: between 55 mV/
decade and 58 mV/decade
MEASURED VALUE
X.XX
[E] TO CONFIRM
7
ADJUSTMENT pH SENSOR
1. Put the pH sensor into the strip
chamber.
2. Screw the cover of the strip
chamber closed hand-tight.
3. Press the [E] key.
RE-INSERT pH-SENSOR
Operation starts.
[E] TO CONFIRM
The following appears on the display if the adjustment failed:
Endress+Hauser
107
Maintenance
TOCII CA72TOC
Step
Display
Description
The adjustment constants are not
accepted.
• Check the buffer solutions.
• Check the sensor or
• Replace the sensor.
Repeat the adjustment.
108
Endress+Hauser
TOCII CA72TOC
Maintenance
8.9
Cleaning
8.9.1
Screen flush
In the version with the pipe back flushing option, water is supplied via the solenoid valve
MV1. This means that, in addition to the sample conditioning system, the pipe is backflushed all the way to the bypass screen.
a0012726
Fig. 46:
Position of the screen
Screen flushing runs automatically after startup, and can be started in three ways:
• Activated manually
• Activated remotely
• Activated automatically
Manual activation of screen flush
Step
Display
Description
1
Endress+Hauser
109
Maintenance
TOCII CA72TOC
Remote activation of screen flush
The screen flush can be triggered using a potential-free contact. Use input 3 of the "binary
in" terminal strip for this purpose.
I
1
1
II
2
1
2
0/4-20
mA
+
0/4-20
mA
+
binary in
2
III
1
IV
binary out
2
40
-
41
-
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
3
4
5
6
7
8
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
RS 232
a0011911
Fig. 47:
1
2
3
4
Signal connection
Calibration, external activation
Adjustment, external activation
Screen flush, external activation
Power flush, external activation
5
6
7
8
Not assigned
Not assigned
Standby, external activation
Channel switchover, external activation (optional)
Automatic activation of screen flush
The screen flush can be triggered automatically.
Select the programming mode for this purpose:
Step
Display
Description
1
2
• For SCREEN FLUSH
n/Day, enter the number of flushes
per day. The factory setting is two
screen flushes per day.
Operation starts automatically on completion of the screen flush process.
110
Endress+Hauser
TOCII CA72TOC
Maintenance
8.9.2
Power flush
During a power flush, the strip chamber and separation chamber are flushed with the
connected pressurized water via solenoid valve MV2.
a0012478
Fig. 48:
Position of the strip and separation chamber
Power flushing runs automatically after startup, and can be started in three ways:
• Activated manually
• Activated remotely
• Activated automatically
Manual activation of power flush
Step
Display
Description
1
Endress+Hauser
111
Maintenance
TOCII CA72TOC
Remote activation of power flush
The power flush can be triggered with a potential-free contact. Use input 4 of the "binary in"
terminal strip for this purpose.
I
1
1
II
2
1
2
0/4-20
mA
+
0/4-20
mA
+
binary in
2
III
1
IV
binary out
2
40
-
41
-
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
3
4
5
6
7
8
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
RS 232
a0011911
Fig. 49:
1
2
3
4
Signal connection
Calibration, external activation
Adjustment, external activation
Screen flush, external activation
Power flush, external activation
5
6
7
8
Not assigned
Not assigned
Standby, external activation
Channel switchover, external activation (optional)
Automatic activation of power flush
The power flush can be triggered automatically.
Select the programming mode for this purpose:
Step
Display
Description
1
2
• For POWER FLUSH
n/Day, enter the number of flushes
per day. The factory setting is two
power flushes per day.
Operation starts automatically on completion of the power flush process.
112
Endress+Hauser
TOCII CA72TOC
Maintenance
8.9.3
Bypass screen
This section describes how the bypass screen is cleaned manually.
a0012489
Fig. 50:
1
2
3
4
Sample conditioning
Upper thread adapter nut
Bypass elbow
Lower thread adapter nut
Bypass screen housing
5
6
7
8
Vent
Bypass screen
O-ring
O-ring
You require the following parts:
• Bottle brush
• Paper towels
Step
Display
Description
1
Endress+Hauser
113
Maintenance
TOCII CA72TOC
Step
2
Display
BYPASS-SCREEN
Description
!
WARNING
There is a risk of infection when
working with wastewater!
Wear protective gloves, goggles
and protective clothing.
‣
MAINTENANCE COMPLETED ?
Clean the bypass screen as follows:
1. Shut off the supply of wastewater
(external sample supply).
[E] TO CONFIRM
2. Set the "Online sample/manual
sample" valve to "Manual sample".
Drain the bypass line. Set the
valve back to the previous
position.
3. Release the lower and upper
thread adapter nut (items 1 +3).
4. Remove the bypass elbow
(item 2).
5. Remove the bypass screen.
6. Clean the bypass screen and the
housing with a brush.
7. Unscreen the vent and open it.
Clean the vent and make sure the
bead is able to move freely.
8. Insert the parts again in the
reverse order.
Make sure that the O-rings
(items 7 and 8) are undamaged
and are correctly positioned.
9. Switch the supply of wastewater
back on.
10. Press the [E] key.
Operation starts.
114
Endress+Hauser
TOCII CA72TOC
Maintenance
8.9.4
Strip and separation chamber
This section describes how to manually clean the strip and separation chamber and replace
the glass bead.
a0012478
Fig. 51:
Position of the strip and separation chamber
You require the following parts:
• 4 mm Allen key
• Injector (part of the scope of supply)
• Glass bead
• Soft brush
• Paper towels to absorb liquid
• Vessel with volume of approx. 150 ml (5 fl.oz) to collect liquid
• Needle-nosed pliers
Endress+Hauser
115
Maintenance
TOCII CA72TOC
a0012343
Fig. 52:
1
2
3
4
5
6
7
8
Strip and separation chamber
pH sensor
Strip chamber cover
Strip chamber thread adapter nut
Coupling
Strip chamber
Coupling (drain)
Separation chamber thread adapter nut
Separation chamber
Step
Display
9
10
11
12
13
14
15
16
17
Separation chamber cover
Magnetic stir bar
Glass bead
Glass frit
O-ring
Coupling (stripping gas connection)
Capillary
Knurled head screw
Seal
Description
1
116
Endress+Hauser
TOCII CA72TOC
Maintenance
Step
2
Display
Description
!
WARNING
There is a risk of infection when
working with wastewater!
Wear protective gloves, goggles
and protective clothing.
‣
Press the [E] key.
3
The strip and separation chamber are
automatically flushed with
pressurized water for 10 s.
4
1. Have a vessel at the ready to
collect the liquid.
2. Release the hose connection of
pump P1 at the strip chamber.
3. Hold the vessel in place and drain
the strip chamber.
4. Absorb any drops of water with
paper towels.
5. Press the [E] key.
5
1. Release the thread adapter nut
(item 3) on the strip chamber.
2. Release the cable of the pH
sensor.
3. Remove the cover (item 2) with
the pH sensor (item 1) from the
strip chamber (item 5).
4. Release the coupling (item 14)
and remove the coupling and the
glass frit (item 12).
5. Release the coupling (item 6) and
take off the hose coupling.
6. Release the thread adapter nut
(item 7) and remove the cover
(item 9).
Endress+Hauser
117
Maintenance
TOCII CA72TOC
Step
Display
Description
7. Use the pliers to remove the magnetic stir bar (item 10) from the separation chamber (item 8).
8. Empty the separation chamber (item 8) with the injector (withdrawn by suction).
9. Connect the empty injector to the suction needle for the sample (connection P2).
10. Using the injector, inject air quickly and force the glass bead out of the bore hole. To make it
easier to remove the glass bead, you can release the coupling (item 4) and release the
separation chamber from the strip chamber.
11. Clean the chambers with a soft brush. In the event of severe fouling, you can separate the strip
chamber and separation chamber from one another by releasing the securing screw. For
complete removal, you must disconnect the connector of the magnetic stirrer controller.
12. Clean the pH sensor (see the Brief Operating Instructions of the pH sensor).
13. Insert a new glass ball.
14. Insert the magnetic stir bar. The thin shaft points upwards.
15. Release the knurled head screw (item 16) and remove the capillary (item 15).
16. Insert a new capillary. Slide in the capillary as far as it will go.
Ensure that the seal (item 17) is seated correctly in the knurled head screw Tighten the
knurled head screw.
17. Fit the hose (P2) onto the capillary.
18. Put the cover on the separation chamber and tighten the thread adapter nut hand-tight.
19. Put the drain pipe onto the coupling and screw the coupling (item 6) closed.
20. Insert the pH sensor and connect the cable.
21. Screw the thread adapter nut hand-tight.
22. Clean the glass frit or replace it. Reinstall the glass frit, the O-ring (item 13) and the coupling.
23. Press the [E] key.
6
1. Connect the hose of pump P1 to
the strip chamber.
2. Press the [E] key.
7
POWER FLUSH
XX
The strip and separation chamber are
automatically flushed with
pressurized water for 180 s.
Measuring operation then starts
automatically.
FILL STRIP AND SEPARATION CHAMBERS
Adjust the pH sensor after cleaning the strip and separation chamber (see the
"Adjusting the pH sensor" section).
118
Endress+Hauser
TOCII CA72TOC
Maintenance
8.9.5
Opening the circuit
This section describes how the dosing head (capillary) is cleaned or replaced.
The furnace temperature is not reduced. The system continues to perform sample
conditioning (stripping).
a0012498
Fig. 53:
Position of the dosing unit
You require the following parts:
• Damp cloth
3
5
4
1
2
a0012492
Fig. 54:
1
2
3
4
5
Endress+Hauser
Dosing
Red screw plug
Capillary
O-ring
O-ring
Dosing head
119
Maintenance
TOCII CA72TOC
Step
Display
Description
1
2
OPENING THE GAS CIRCUIT
Clean the needle as follows:
1. Remove the hose P2 from the
capillary of the dosing head.
MAINTENANCE COMPLETED ?
[E] TO CONFIRM
2. Release the red screw plug (item
1).
3. Remove the dosing head (item 5).
4. Using a damp cloth, remove the
salt residue on the capillary or
replace the capillary.
When installing the
capillary (item 2), ensure
that the capillary is
protruding from the bottom
of the dosing head by
approx. 10 mm (0.4 ").
5. Check the O-ring (item 4), and
also check the O-ring (item 3) if
replacing the capillary.
6. Insert the dosing head and
tighten the red screw plug.
7. Fit the hose P2 onto the capillary.
8. Press the [E] key.
Operation starts.
120
Endress+Hauser
TOCII CA72TOC
Maintenance
8.9.6
Combustion pipe
This section describes how the combustion pipe is cleaned and replaced, or how the entire
furnace is replaced.
The furnace heating system is switched off.
a0012530
Fig. 55:
Position of the furnace
You require the following parts:
• Auxiliary tool for the combustion pipe insert
• Crucible tongs
• Heat-resistant gloves
• Dust mask FFP3
Endress+Hauser
121
Maintenance
TOCII CA72TOC
a0012387
Fig. 56:
Furnace
1
Dosing head
6
2
3
4
5
Red screw plug with O-ring
Furnace cover
Thread adapter nut
Furnace outlet with O-ring
7
8
9
10
Step
1
Display
Knurled head screw with support and sealing
ring
Clamp screw
Optional heated salt trap
Safety guard
Combustion pipe
Description
!
CAUTION
Inhalation of catalyst dust poses a
health hazard.
Wear a breathing mask when
replacing the catalyst.
‣
122
Endress+Hauser
TOCII CA72TOC
Maintenance
Step
2
Display
Description
!
COMBUSTION PIPE
CAUTION
Contact with hot parts of the
combustion furnace may cause
injury!
Use heat-resistant gloves.
‣
REMOVE HOSE P2 FROM FURNACE
The power supply for the combustion
furnace is switched off. The furnace
cools down.
1. Release the hose at the dosing
head (item 1).
2. Press the [E] key.
[E] TO CONFIRM
3
!
COMBUSTION PIPE
TEMPERATURE
OPEN FURNACE
[E] TO CONFIRM
=
XXX
CAUTION
Contact with hot parts of the
combustion furnace may cause
injury!
Use heat-resistant gloves.
‣
°C
If the insert in the combustion
pipe is removed when very hot
(over 300 °C), cracks may form
in the insert and the
combustion pipe if they cool
down too quickly. This results in
a higher basic line and has a
negative impact on the
operation of the measuring
device.
1. If a heated salt trap is installed,
release the hose connection and
the electrical collection at the
heated salt trap.
2. Remove the safety guard.
3. Swivel the furnace outwards.
4. Pull the heated salt trap down
from the furnace outlet by
turning it gently back and forth.
5. Swivel the furnace back in and
lock it.
6. When the temperature has
dropped to approx.
300 °C, slacken the lower knurled
head screw (item 6).
7. Allow the furnace to cool to
below 50 °C (not necessary if
replacing the furnace).
8. Press the [E] key.
If you are replacing the entire furnace,
proceed with Step 4a.
If you are cleaning or replacing the
combustion pipe, proceed with Step
4b.
Endress+Hauser
123
Maintenance
TOCII CA72TOC
Step
4a
Display
Description
Only if replacing the furnace:
!
CAUTION
Contact with hot parts of the
combustion furnace may cause
injury!
Use heat-resistant gloves.
COMBUSTION PIPE
‣
TEMPERATURE
=
XXX
MAINTENANCE COMPLETED?
[E] TO CONFIRM
°C
1. Detach the outflow hose on the
mounting plate or on the salt
trap, if a heated salt trap
(optional) is fitted.
2. Release the locking button on the
furnace and swivel the furnace
outwards.
3. Release the electrical plug-in
connections between the
analyzer and the furnace.
4. Remove the furnace.
5. Press the [E] key.
Continue with Step 7.
124
Endress+Hauser
TOCII CA72TOC
Maintenance
Step
4b
Display
Description
Only if cleaning or replacing the combustion pipe:
!
CAUTION
Contact with hot parts of the
combustion furnace may cause
injury!
Use heat-resistant gloves.
COMBUSTION PIPE
‣
TEMPERATURE
=
XXX
MAINTENANCE COMPLETED?
[E] TO CONFIRM
°C
1. If a heated salt trap is not fitted,
release the hose connection
between the furnace outlet (item
5) and the mounting plate.
2. Remove the knurled head screw
(item 6). Remove the furnace
outlet and the O-ring from the
combustion pipe.
3. Release the red screw plug (item
2) and remove the dosing head
(item 1).
4. Remove the safety guard.
5. Release the thread adapter nut on
the furnace entry (item 4) and
remove the furnace cover.
6. Remove the O-ring and support
ring.
7. Unlock the furnace and pull it out.
8. Using the auxiliary tool, pull the
combustion pipe insert
approx. 10 mm (0.4") out of the
combustion pipe. Then remove
the combustion pipe insert using
the crucible tongs, and take out
the filling.
Put the waste in a container
for inorganic materials.
Dispose of the used catalyst
as hazardous waste in
accordance with local laws
and safety regulations. Do
not pour it down the drain
or into a garbage can!
9.Lift the combustion pipe
under the furnace and
using the crucible tongs remove it
from the furnace from the top.
10. If necessary, clean the
combustion pipe using a brush.
Endress+Hauser
125
Maintenance
TOCII CA72TOC
Step
5
Display
Description
Reassembling the combustion unit
1. Insert the combustion pipe into the tube furnace.
2. Place the combustion pipe insert with filler (approx. 25 g high-temperature catalyst) into the
combustion pipe.
3. Check and clean the support ring and sealing ring for the furnace cover and insert the sealing
ring and the support ring.
4. Fit the cleaned furnace cover and the thread adapter nut on the furnace entry and tighten the
thread adapter nut.
5. Mount the dosing head with the O-ring, and tighten the red screw plug.
6. If you are not using the optional heated salt trap, fit the furnace outlet with a glass fiber fabric
as the salt trap. Roll two fabrics together loosely and put the fabrics loosely into the furnace
outlet. Approx. 10 mm (0.4") must remain free at the top end to catch the salt.
7. If you are using the optional heated salt trap, leave the furnace outlet empty.
8. Insert the cleaned furnace outlet and a clean O-ring into the combustion pipe and tighten the
knurled head screw hand-tight.
9. If you are not using the optional heated salt trap, connect the hose from the furnace outlet to
the bulkhead gland on the mounting plate.
6
If you are using the optional heatable salt trap:
1. Turn the salt trap gently to push it onto the furnace outlet nozzle.
2. Make sure that the seal seals the glass nozzle with a slight suction effect. You can adjust the
seal with the clamp screw. The seal should not be too tight however.
3. Push the salt trap under the furnace.
4. Fold down the retaining bar and rest the filter on the bar.
5. Plug in the electrical contact and lock it.
6. Fit the hose on the heatable salt trap and screw it down.
Press the [E] key.
7
1. Once you have replaced the entire
furnace, connect the hose to the
mounting plate or to the heated
salt trap (optional). Connect the
electrical plug-in connections
between the furnace and the
analyzer.
2. Move the furnace into position
and secure it.
3. Press the [E] key.
126
Endress+Hauser
TOCII CA72TOC
Maintenance
Step
8
Display
Description
1. Install the hose at the injection
unit.
2. Check that the hose is correctly
inserted in the magnetic valve 8.
3. Press the [E] key.
9
The carrier gas valve MV7 (relay #7)
is opened as soon as 85% of the set
temperature is reached.
The system flushes permanently with
carrier gas.
Sample conditioning takes place (strip
chamber) after the furnace is being
heated.
Operation starts automatically.
Now carry out a leakage test (see
"Leakage test" section).
Endress+Hauser
127
Maintenance
TOCII CA72TOC
8.9.7
Leakage test
This section describes how the gas circuit is checked for leaks.
Check the leak-tightness of the gas circuit after each modification to the furnace.
You require the following parts:
• Drain plug from the accessories supplied
1
2
a0012531
Fig. 57:
1
2
Left-hand side of the housing
Gas outlet
Switch for membrane compressor
Step
1
Display
Description
S E R V I C E
PUMPS
CALIBRATION
CLEANING
FILTERS
2
SCREEN FLUSH
POWER-FLUSH
BYPASS-SCREEN
STRIPPER+SEPARATION
OPEN GAS CICUIT
COMBUSTION PIPE
LEAKAGE TEST
LEAKAGE TEST
1. Switch off the membrane
compressor (item 2).
2. Seal the gas outlet with a plug
(left-hand side panel, item.1).
SWITCH MEMBRANE COMPRESSOR OFF
3. Press the [E] key.
SHUT GAS OUTLET
[E] TO CONFIRM
128
Endress+Hauser
TOCII CA72TOC
Maintenance
Step
Display
Description
3
Press the [->] key.
The carrier gas valve is opened and
pressure is applied to the gas circuit.
The pressure is shown on the display.
The carrier gas valve is closed
automatically:
• If the pressure has exceeded 100
mbar or
• Once 7 seconds have elapsed.
If the pressure does not reach
100 mbar, this indicates a
sizable leak.
4
LEAKAGE TEST
CARRIER VALVE SHUT
x.xxx mbar/min
XXX mbar
XX
[E] TO CONFIRM
sec
The pressure loss rate is displayed
(mbar/min) after 30 s.
The pressure loss must be < 3 mbar/
min. Values are typically between 0.5 and -2.0 mbar/min.
If the loss in pressure is > 3 mbar/
min, divide the gas circuit into smaller
sections. Bridge the individual
components using a hose, and repeat
the leakage test until you have found
the leak. If there is no pressure loss in
a bridged component e.g. a
component of the furnace with the
salt trap, the leak is in the bridged
component.
Press the [E] key.
5
LEAKAGE TEST
1. Remove the plug from the gas
outlet (left-hand side panel).
2. Switch on the membrane
compressor.
OPEN GAS OUTLET
SWITCH MEMBRANE COMPRESSOR ON
3. Press the [E] key.
Operation starts.
[E] TO CONFIRM
Endress+Hauser
129
Maintenance
TOCII CA72TOC
Possible points where a leak might be present:
• Furnace seals
• Acid filter seal at the glass
• Condensate drain
• Water block filter
To find the leak, the following tools in the "Maintenance tool kit" (see "Spare parts" section)
are available for the purpose of bridging components:
• Hose D 3/5 mm FPM
• Hose connector 1/8 - 1/8 PP
To seal the gas outlet at the furnace:
• Protective cap
• Reducing adapter 8/4 mm, straight
To seal the gas outlet at the cabinet:
• Locking cap M3 EPDM
130
Endress+Hauser
TOCII CA72TOC
Maintenance
8.10
Filter
8.10.1
Replacing the acid filter
This section describes how the acid filter is replaced.
Replace the acid filter:
• If it is blocked or worn out. This is noticeable from the flow rate and pressure level of the
gas circuit.
• If zinc or copper become completely and obviously discolored.
a0012500
Fig. 58:
Position of the acid filter
You require the following parts (part of the wear parts kit):
• Glass fiber fabric
• Zinc pellets
• Copper powder
1
5
5
5
6
6
6
2
3
7
4
8
a0012316
Fig. 59:
1
2
3
4
Endress+Hauser
Acid filter
Coupling
Retaining clip
Filter glass body
GL coupling
5
6
7
8
Glass fiber fabric
Zinc
Copper
Glass fiber fabric
131
Maintenance
TOCII CA72TOC
Step
Display
Description
1
2
REPLACE ACID FILTER
!
CAUTION
Risk of injury due to aggressive
materials.
Wear protective gloves!
‣
MAINTENANCE COMPLETED ?
1. Release the couplings (items 1
and 4).
2. Remove the acid filter (item 3)
from the retaining clips (item 2).
3. Remove the filling in the acid
filter.
4. Clean the glass if necessary.
[E] TO CONFIRM
5. Roll the fabric into a roll and push
it into the filter glass (item 8).
Do not press it in too hard.
Shorten the fabric if necessary.
6. Fill the glass up to the halfway
mark with copper and then with
zinc (see order of items 6 and 7).
Allow enough room for the
second piece of fabric.
7. Roll the fabric into a roll and use
it to close the filling in the acid
filter (item 5).
8. Clean the O-rings and seal the
acid filter. Make sure that the
fabric does not extend as far as
the plug to ensure the filter
housing is sealed properly (see
zoom detail of å 59).
9. Fit the acid filter into the
retaining clips and connect the
filter.
10. Press the [E] key.
Operation starts (initially without a
measured value).
132
Endress+Hauser
TOCII CA72TOC
Maintenance
8.10.2
Replacing the gas filter
This section describes how the gas filter is replaced.
Replace the gas filter if it is blocked.
a0012500
Fig. 60:
Position of the gas filter
1
IN
2
3
OUT
a0012307
Fig. 61:
1
2
3
Endress+Hauser
Gas filter
Coupling
Gas filter
Coupling
133
Maintenance
TOCII CA72TOC
Step
Display
Description
1
2
REPLACE GAS FILTERS
1. Release the couplings (items 1
and 3).
2. Replace the gas filter (item 2).
Take the direction of flow into
account.
MAINTENANCE COMPLETED ?
3. First secure the filter in the
coupling (item 3) and then
connect the filter to the acid filter
with the coupling (item 1). Make
sure that the cone is properly
positioned on the filter.
[E] TO CONFIRM
4. Tighten the two couplings.
Operation starts (initially without a
measured value).
134
Endress+Hauser
TOCII CA72TOC
Maintenance
8.10.3
Replacing the O2 filter
This section describes how the O2 filter is replaced.
The gas connection block with the O2 filter is mounted on the rear of the analyzer.
You require the following parts:
• Open-ended wrench
• Needle-nosed pliers
1
2
3
a0012321
Fig. 62:
1
2
3
Gas connection block with O2 filter
Gas connection block
O2 filter
Coupling
Step
Display
Description
1
2
!
CAUTION
Risk of injury due to release of
pressure
Wear protective goggles!
‣
1. Close the valve for the carrier gas
supply.
2. Relieve the pressure on the
pressure line before opening the
hose connection to prevent injury
from the uncontrolled release of
pressure.
3. Press the [E] key.
Endress+Hauser
135
Maintenance
TOCII CA72TOC
Step
3
Display
Description
1. Release the coupling (item 3).
REPLACE O2 FILTER
2. Check the O2 filter (item 2) for
possible signs of wear. Replace
the filter if necessary.
3. Screw the coupling into the gas
connection block.
MAINTENANCE COMPLETED ?
4. Press the [E] key.
[E] TO CONFIRM
4
1. Reconnect the gas supply.
2. Open the gas supply valve.
3. Press the [E] key.
5
PLEASE WAIT
6
The analyzer display switches to step
6 after 10 seconds.
REPLACE O2 FILTER
XX
The furnace heats up. Measuring
operation starts automatically as soon
as 90% of the set temperature is
reached and the CO2 value has
dropped below the threshold value.
During this time, sample conditioning
takes place (strip chamber) and pH
regulation is activated
Measuring operation commences if
these two conditions are met.
136
Endress+Hauser
TOCII CA72TOC
Maintenance
8.10.4
Replacing the heated salt trap
This section describes how to replace the heated salt trap (referred to as "heated filter" in the
menu).
a0012515
Fig. 63:
Position of heated salt trap
You require the following parts:
• 4 mm Allen key
• Deionized water
• Heat-resistant gloves
3
3
4
5
6
a0012516
Fig. 64:
1
2
7
Endress+Hauser
Heated salt trap when installed
Retaining bracket
Heated salt trap with insulation
Clamp screw
a0012517
Fig. 65:
3
4
5
6
Heated salt trap, open
Allen screws
Filter housing, top section
Filter cartridge
O-ring
137
Maintenance
TOCII CA72TOC
To ensure that the furnace does not cool down too much during the cleaning process, it
is heated between the point the salt trap is removed and the salt trap is installed.
Extended measuring device downtime results if the furnace cools down too much, and
should therefore be avoided.
Step
Display
Description
1
2
!
CAUTION
Contact with hot parts of the
combustion furnace may cause
injury!
Use heat-resistant gloves.
‣
Carry out the following tasks quickly
to ensure the furnace does not cool
down too much:
1. Release the hose connections on
the dosing head.
2. Unlock the furnace and swivel it
outwards.
3. Release the electrical connection
to the salt trap.
4. Press the [E] key.
3
Confirm that you disconnected the
electrical connection to the salt trap.
The furnace is reheated and the
temperature displayed.
138
Endress+Hauser
TOCII CA72TOC
Maintenance
Step
4
Display
Description
1. Release the hose at the outlet of
the heated salt trap.
2. Raise the salt trap (item 2)
slightly.
3. Fold back the retaining bracket
(item 1).
4. Remove the salt trap (item 2)
from below.
5. Remove the insulation.
6. Release the two screws (item 3)
and remove the top part of the
filter housing (item 4).
7. Clean the inside of filter cartridge
(item 5) and the filter housing
with deionized water.
8. Clean the O-ring (item 6) and fit
it back into the groove.
9. Screw the filter housing together
and insert it into the insulation.
10. Press the [E] key.
5
Carry out the following tasks quickly
to ensure the furnace does not cool
down too much:
1. Fit the salt trap on the glass
nozzle of the furnace.
2. Make sure that the seal seals the
glass nozzle with a slight suction
effect. You can adjust the seal
with the clamp screw (item 7).
The seal should not be too tight
however.
3. Push the salt trap under the
furnace.
4. Fold down the retaining bar and
rest the salt trap on the bar.
5. Reestablish the electrical
connection.
6. Press the [E] key.
6
The furnace is reheated and the
temperature displayed.
1. Connect the hose to the outlet of
the salt trap.
2. Swivel the furnace back and
ensure that the hose passes
comfortably through the rear
panel and is not buckled.
3. Lock the furnace.
4. Reestablish the hose connection
at the dosing head.
5. Press the [E] key.
Endress+Hauser
139
Maintenance
TOCII CA72TOC
Step
140
Display
Description
7
The analyzer waits until the
temperature is 30 ° below the set
temperature.
8
Press the [E] key to start measuring
operation.
Endress+Hauser
TOCII CA72TOC
Maintenance
8.10.5
Replacing the filter pads in the ventilators
This section describes how the filter pads in the ventilators are checked and replaced.
a0021275
Fig. 66:
Position of filter pads
To replace the filter pads, you will require the following parts:
• Replacement filter pad AM 115P (2x)
• Replacement filter pad AM 335P (1x)
Proceed as follows:
Endress+Hauser
1.
Remove the safety guard (no tool necessary).
2.
Check if the filter pads are soiled.
3.
Replace the soiled filter pads.
4.
Put the safety guard back in.
Ensure that the ventilation slits on the safety guard are pointing down.
141
Repair
TOCII CA72TOC
9
Repair
9.1
Spare parts
pH electrode for strip chamber
Order No.: CPS71-1TB2GSA
Wear parts kit for one year
Order No: 71095156 for versions with standard furnaces
142
Name
Quantity
Flange seal D 53.9 x 63 x 3.1 mm
1
O-ring D 36.09 x 3.53 mm FKM
1
O-ring D 1.5 x 1.5 mm FPM
2
O-ring D 8 x 2 mm FPM
1
Capillary 1.6 x 0.8 x 71 mm AL23
1
Catalyst 800 - 900 °C 75 g pack
1
Cone for coupling 1/8" 28 UNF ETFE
2
Coupling with nozzle 1/4"-28 D1.55 ETFE
4
Coupling with nozzle 1/4"-28 D3.20 ETFE
2
Fiberglass precut part, acid filter 24 x 230 mm
8
Copper, powder 50g pack
2
Zinc, granules, 60g pack
2
O-ring D 33 x 3 mm EPDM
1
O-ring D 12 x 1.5 mm EPDM
1
Filter cartridge 12 x 1 mm frit, porosity 4
1
O-ring D 35 x 3 mm EPDM
1
Capillary 1.6 x 0.8 x 45 mm AL23
1
Membrane for solenoid valve 330 A EPDM
2
Hose cassette, Ismatec CA71
2
Hose 2.79 / 4.59 PharMed violet/white
4
Hose 0.76 / 2.36 PharMed black/black
12
Glass balls D 2 mm 10-piece pack
1
Replacement filter pad AM 115P, air outlet
2
Replacement filter pad AM 335P, air outlet
1
Replacement membrane filter (gas filter)
1
Dust mask FFP3
3
Endress+Hauser
TOCII CA72TOC
Repair
Name
Quantity
Locking cap M3 EPDM
1
Protective cap
1
Order no.: 71095149 for versions with heated salt trap
Contains all the parts of 71095156 and also:
Name
Quantity
O-ring D 64 x 2.5 mm FPM
1
Seal, glass tube heating filter
1
Filter sleeve, 25 x 64 20μ PTFE
2
Coupling D 4/6 G1/8 PP
1
Maintenance tool kit
Order No.: 71102317
Endress+Hauser
Name
Quantity
Locking cap M3 EPDM
1
Hose D 3/5 mm FPM
0.6
Protective cap
1
Hose connector 1/8 - 1/8 PP
2
Dosing needle D 0.84 mm, green
1
Injector 20 ml
1
Hose iD 1/2 mm Tygon 400 mm
1
Auxiliary tool for the salt trap
1
Crucible tongs 200 mm in length
1
Heat-resistant gloves (1 pair)
1
Gloves, resistant to acids and bases, large (1 pair)
1
Installation tool for seal set
1
Reducing adapter 8/4 mm, straight
1
143
Repair
TOCII CA72TOC
Wear parts kit for strip and separation chamber
Order No.: 71101606
Name
Quantity
O-ring D 1.5 x 1.5 mm FPM
1
O-ring D 8 x 2.5 mm FPM
1
O-ring D 33 x 3 mm EPDM
1
O-ring D 12 x 1.5 mm EPDM
1
Filter cartridge 12 x 1 mm frit, porosity 4
1
O-ring D 35 x 3 mm EPDM
1
Capillary 1.6 x 0.8 x 45 mm AL23
2
Glass balls D 2 mm 10-piece pack
1
Stirrer support, complete, for 100 μ particles
1
Wear parts kit for acid filter
Order No.: 71101607
Name
Quantity
Zinc, granules, 60g pack
3
Copper, powder 50g pack
3
Fiberglass precut part, acid filter 24 x 230 mm
2
Wear parts kit for predilution pump
Order No.: 71101608
144
Name
Quantity
Coupling, short thread 1/8" - 28 UNF ETFE
1
Cone for coupling 1/8" - 28 UNF ETFE
1
Seal set D10
1
Endress+Hauser
TOCII CA72TOC
Repair
9.2
!
Decommissioning
CAUTION
There is a risk of infection if you come into contact with wastewater.
‣ Wear protective gloves, protective goggles and a protective laboratory coat!
Proceed as follows to take the analyzer out of service:
1.
Switch off the wastewater pump.
2.
If sample preparation is present: Trigger a screen flush (CLEANING - SCREEN FLUSH).
Allow the bypass to drain.
3.
Optional dilution water pump P5: Flush the pump with 5 % acid followed by DI water.
Use the TEST menu to control pump P5. You can pump the acid and the DI water directly
out of the feeder tank.
4.
Flush the hoses.
Set valve 1 to Manual sample and place a container with deionized water under the
valve.
TEST - TEST OF OUTPUTS - PUMPS
Enter 400% for pump P1 and pump P4 (optional) and let the pumps pump for some
time.
Remove the acid hose of pump P3 from the acid canister and insert it into a container
with deionized water. Also let this pump run at 400 % for some time.
5.
Clean the vessels
Trigger an automatic flush of the strip chamber (CLEANING - POWER FLUSH).
Afterwards, perform manual cleaning of the strip and separation chamber (CLEANING
- STRIPPER & SEPARATION). The pH sensor must be stored wet. Fill the protection cap
with 3-mole KCl solution and fit the protection cap onto the sensor.
6.
Empty the hoses
Open the hose cassettes of pumps P1, P2, P3 and (if optional dilution takes place) P4.
Let the rinse water run out of the hoses.
Remove the canister with the standard.
TEST - TEST OF OUTPUTS - SWITCH OUTPUTS
Switch the switch outputs SA1 and SA4 to "ON". Wait until the pipes for standard 1 and
2 are empty.
Switch the switch outputs "OFF" again.
Remove the feeder tank.
7.
Switch off the main switch on the analyzer.
8.
Empty the combustion pipe insert
Empty the combustion pipe insert (see "Maintenance - Combustion pipe" section).
Do not dispatch the analyzer with the combustion pipe insert still inserted (risk of
breakage).
Empty the glass insert.
Assemble the combustion unit.
9.
Remove the exhaust (if present).
10. Close off the carrier gas supply.
Relieve the pressure on the pressure line before opening the hose connection to prevent
injury from the uncontrolled release of pressure.
Unscrew the hose for the carrier gas supply on the left side panel.
Detach the hose on the pressure-reducing valve of the carrier gas cylinder or gas
preparation system.
Endress+Hauser
145
Repair
TOCII CA72TOC
9.3
Return
The product must be returned if repairs or a factory calibration are required, or if the wrong
product was ordered or delivered. Due to legal specifications and as an ISO-certified
company, Endress+Hauser is required to follow certain procedures when handling all
returned products that have been in contact with medium.
To ensure that your device is returned swiftly, safely and in an appropriate manner,
please read the procedures and conditions on the Endress+Hauser website at
www.services.endress.com/return-material
9.4
Disposal
The device contains electronic components and must therefore be disposed of in accordance
with regulations on the disposal of electronic waste.
Please observe local regulations.
146
Endress+Hauser
TOCII CA72TOC
Accessories
10
Accessories
Reagents
‣ CAY450-V10AAE, 1000 ml stripper reagent für CA72TOC
‣ CAY451-V10C01AAH, 1000 ml parent solution 5,000 mg/l TOC
‣ CAY451-V10C10AAH, 1000 ml parent solution 100,000 mg/l TOC
Technical buffer solutions, accuracy 0.02 pH, according to NIST/DIN
‣ ph 4.00; 20 x 18 ml (0.68 fl. oz); order no. CPY20-C01A1
‣ ph 7.00; 20 x 18 ml (0.68 fl. oz); order no. CPY20-E01A1
Furnace conditioning station
230 V AC 50/60 Hz order number: 71103492
115 V AC 50/60 Hz order number: 71103493
‣
‣
Standard furnace
To be used as a second furnace or as a substitute
Cannot be used with heated salt trap
Order number: 71091188
‣
‣
‣
Furnace for heated salt trap
To be used as a second furnace or as a substitute
Order number: 71103888
‣
‣
Kit CA72TOC heated salt trap
As replacement for maintenance (decreases the maintenance time) or as replacement
Order number: 71101532
‣
‣
Endress+Hauser
147
Technical Data
TOCII CA72TOC
11
Technical Data
11.1
Input
Measured variable
TOC (TC possible without stripping unit)
Measuring range
Version
Measuring range
A
0.25 to 600 mg TOC/l
B
1 to 2400 mg TOC/l
C
2.5 to 6000 mg TOC/l
D
5 to 12000 mg TOC/l
With optional predilution, the measuring range can be expanded by a factor of 20.
Signal inputs
8 signal inputs 24 V DC, active, load max. 500
Input #1
Service, trigger calibration
Input #2
Service, trigger adjustment
Input #3
Service, trigger screen flush
Input #4
Service, trigger power flush
Input #5
Input #6
11.2
Input #7
Trigger standby
Input #8
Trigger channel switchover (optional)
Output
Output signal
Measuring channel 1: 0/4 to 20 mA, galvanically isolated
Measuring channel 2 (for two-channel version): 0/4 to 20 mA, galvanically isolated
Load: max. 500
Data interface
RS 232 C, proprietary, for data output and remote operation (optional)
Alarms
4 outputs: Limit value alarm, fault message, standby message and operating check,
potential-free, normally closed (max. 0.25 A / 50 V)
11.3
Power supply
Supply voltage
115/230 V AC, 50/60 Hz, 800 VA
Fuses
Assembly
Fuses
Power distribution
2.5 A, slow-blow, design: fine-wire fuse 6.3 x
32
Relay
4 A per relay, slow-blow, design: TR5
Power unit
2 A, slow-blow, design: fine-wire fuse 5 x 20
TOCII CA72TOC is suitable for connection to industrial power supply systems in accordance with EN
61326-1, Class A.
148
Endress+Hauser
TOCII CA72TOC
Technical Data
11.4
Performance characteristics
In accordance with ISO 15839
The performance characteristics have been determined in accordance with ISO 15839, Annex B. 300 μl sample
were metered into the TOCII (CA72TOC-B1A0B1) per measurement. This resulted in a measuring range from 4
to 800 mg/l. The following data refer to this device.
Slight deviations should be factored in if applying the performance characteristics to other measuring ranges.
LOD (limit of detection)
0.75 % of end of measuring
range
LOQ (limit of quantitation)
2.5 % of end of measuring
range
Short-term drift
0.5 %/day
Repeatability precision at 20 % of the measuring range
0.4 %
Resolution limit at 20 % of the measuring range (LDC)
1.1 %
Systematic measured value deviation at 20 % of the measuring range (BIAS)
0.4 %
Repeatability precision at 80 % of the measuring range
1.6 %
Resolution limit at 80 % of the measuring range (LDC)
4.6 %
Systematic measured value deviation at 80 % of the measuring range (BIAS)
2.4 %
11.5
Ambient temperature
range
5 to 35 °C (41 to 95 °F)
Humidity
20 to 80 %, non-condensing
Degree of protection
IP 54
11.6
Process
Sample temperature
4 to 40 °C (39 to 104 °F)
Sample flow rate
20 ml/min (0.32 US gal/hr)
Sample feeder volume
90 ml (3 fl.oz)
PA-2, PA-3 and PA-9 are available for selection as sample preparations.
Sample consistency
Aqueous
Flammable substances should not occur in combustible concentrations - sample
dilution is then necessary.
Sample supply
Non-pressurized infeed to the analyzer from sample preparation.
11.7
Endress+Hauser
Environment
Mechanical construction
Weight
Approx. 75 kg (165 lbs)
Material
Housing
Aluminum, powder-coated (RAL7035)
Window
Glass, conductive coating
Valve seals
EPDM, FPM, FFKM
Pump hoses
Ismapren
Pumps and pump seals
PTFE, FFKM
Reagent and sample hoses
PTFE, PE
Exhaust gas and ventilation hoses
PTFE, PE
Outflow hoses
PTFE
149
TOCII CA72TOC
Index
A
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
Adjusting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Adjusting pump P2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Adjusting the analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Adjusting the pH sensor . . . . . . . . . . . . . . . . . . . . . . . . . 105
ALARM LIMITS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
ALARM RECORD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
B
BASIC DATA. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Bypass screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
C
Calibrating the analyzer . . . . . . . . . . . . . . . . . . . . . . . . . 100
Check
Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Installation and function. . . . . . . . . . . . . . . . . . . . . . . 34
Citric acid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69, 109
Combustion pipe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
COMPLETE RECORDS . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Compressed air and water supply . . . . . . . . . . . . . . . . . . 10
Connection
Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Distributor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
H
Hose replacement
P1 and P4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
P2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
P3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
I
Incoming acceptance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9, 19
Installation conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
K
Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
L
Leakage test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
LISTS
ALARM RECORD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
COMPLETE RECORDS . . . . . . . . . . . . . . . . . . . . . . . . . . 42
MAINTENANCE RECORD. . . . . . . . . . . . . . . . . . . . . . . 45
MIN MAX AVERAGE . . . . . . . . . . . . . . . . . . . . . . . . . . 47
RECORD ALL DATA . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
M
Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131
Filter pads. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Adjusting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Adjusting the analyzer . . . . . . . . . . . . . . . . . . . . . . . . . 97
Adjusting the pH sensor . . . . . . . . . . . . . . . . . . . . . . 105
Bypass screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Calibrating the analyzer . . . . . . . . . . . . . . . . . . . . . . 100
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Combustion pipe . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
Empty volume dosing . . . . . . . . . . . . . . . . . . . . . . . . 102
Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131
Leakage test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
Opening the circuit . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Power flush . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Pumps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Replacing the acid filter . . . . . . . . . . . . . . . . . . . . . . 131
Replacing the filter pads. . . . . . . . . . . . . . . . . . . . . . 141
Replacing the gas filter . . . . . . . . . . . . . . . . . . . . . . . 133
Replacing the heated filter . . . . . . . . . . . . . . . . . . . . 137
Replacing the O2 filter . . . . . . . . . . . . . . . . . . . . . . . 135
Screen flush . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Service software. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Strip and separation chamber . . . . . . . . . . . . . . . . . 115
Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
MAINTENANCE RECORD. . . . . . . . . . . . . . . . . . . . . . . . . . 45
Measuring range adjustment . . . . . . . . . . . . . . . . . . . . . . 55
MEASURING SITE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Methods of installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
MIN MAX AVERAGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
G
N
D
Declaration of conformity. . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Decommissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
Designated use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Device designation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Diagnostics and troubleshooting . . . . . . . . . . . . . . . . . . . 60
Dilution water pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
Distributor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Dosing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
E
Electrical technician . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Empty volume dosing . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
Error messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
F
Gas flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
150
Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Endress+Hauser
TOCII CA72TOC
O
Opening the circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Operating elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Optimization. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
P
Personnel requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Pivoting frame . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Post-connection check . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Potassium hydrogen phthalate . . . . . . . . . . . . . . . . . . . . 16
Power flush. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Power unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Preparing chemicals
Chemicals preparation . . . . . . . . . . . . . . . . . . . . . . . . 15
Product safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Programming mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
T
Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148
TEST
INPUT TEST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
OUTPUT TEST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Two-channel
External changeover . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Time-controlled changeover . . . . . . . . . . . . . . . . . . . . 53
U
Updating the software . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
W
Wall mounted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Wiring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Workplace safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
R
RANGE DATA. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Reagent quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Recording mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
Replacing the acid filter . . . . . . . . . . . . . . . . . . . . . . . . . 131
Replacing the gas filter . . . . . . . . . . . . . . . . . . . . . . . . . . 133
Replacing the heated filter . . . . . . . . . . . . . . . . . . . . . . . 137
Replacing the O2 filter . . . . . . . . . . . . . . . . . . . . . . . . . . 135
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
S
Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Salt load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Screen flush . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Set brightness. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
SET CLOCK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Set contrast. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
SETTING
ALARM LIMITS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
BASIC DATA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
MEASURING SITE . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
RANGE DATA. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
SET BRIGHTN./CONTR.. . . . . . . . . . . . . . . . . . . . . . . . 41
SET CLOCK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Signal connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
Spot view. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Stand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Standard
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Topping up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Standard concentrations. . . . . . . . . . . . . . . . . . . . . . . . . . 15
Standby mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Strip and separation chamber . . . . . . . . . . . . . . . . . . . . 115
Stripping reagent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Endress+Hauser
151
www.addresses.endress.com