Teledyne Analytical Instruments OPERATING INSTRUCTIONS Model 6600 Oil in Water Analyzer
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Oil in Water Analyzer
OPERATING INSTRUCTIONS
Model 6600
Oil in Water Analyzer
DANGER
HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM.
PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.
HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST FOR A
TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.
ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING. BEFORE
CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/MANAGER.
Teledyne Analytical Instruments
P/N M71055
12/22/99
ECO # 99-0000
i
Model 6600
Copyright © 1999 Teledyne Analytical Instruments
All Rights Reserved. No part of this manual may be reproduced, transmitted,
transcribed, stored in a retrieval system, or translated into any other language or computer
language in whole or in part, in any form or by any means, whether it be electronic,
mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of
Teledyne Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 917491580.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us
free from defects of material and of construction, and that our liability shall be limited to
replacing or repairing at our factory (without charge, except for transportation), or at
customer plant at our option, any material or construction in which defects become
apparent within one year from the date of shipment, except in cases where quotations or
acknowledgements provide for a shorter period. Components manufactured by others bear
the warranty of their manufacturer. This warranty does not cover defects caused by wear,
accident, misuse, neglect or repairs other than those performed by Teledyne or an authorized service center. We assume no liability for direct or indirect damages of any kind and
the purchaser by the acceptance of the equipment will assume all liability for any damage
which may result from its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our
apparatus, and to make any alterations in the dimensions, shape or weight of any parts, in
so far as such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its user, and serves as a tool by
which valuable data can be gathered. The information provided by the instrument may
assist the user in eliminating potential hazards caused by his process; however, it is
essential that all personnel involved in the use of the instrument or its interface, with the
process being measured, be properly trained in the process itself, as well as all instrumentation related to it.
The safety of personnel is ultimately the responsibility of those who control process
conditions. While this instrument may be able to provide early warning of imminent
danger, it has no control over process conditions, and it can be misused. In particular, any
alarm or control systems installed must be tested and understood, both as to how they
operate and as to how they can be defeated. Any safeguards required such as locks, labels,
or redundancy, must be provided by the user or specifically requested of Teledyne at the
time the order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions. The
purchaser is responsible for the training of personnel, for providing hazard warning
methods and instrumentation per the appropriate standards, and for ensuring that hazard
warning devices and instrumentation are maintained and operated properly.
Teledyne Analytical Instruments, the manufacturer of this instrument, cannot
accept responsibility for conditions beyond its knowledge and control. No statement
expressed or implied by this document or any information disseminated by the manufacturer or its agents, is to be construed as a warranty of adequate safety control under the
user’s process conditions.
ii
Teledyne Analytical Instruments
Oil in Water Analyzer
Table of Contents
Part I: Control Section ................................. Part I
Part II: Analysis Section ............................. Part II
Part III: Oil in Water Sample System ......... Part III
Appendix, Generic Info................................... A-1
Teledyne Analytical Instruments
iii
Model 6600
iv
Teledyne Analytical Instruments
Part I: Control Section
OPERATING INSTRUCTIONS
Model 6600
Oil in Water Analyzer
Part I: Control Section
of the Control/Analysis Unit
Z-PURGED
CLASS I, DIVISION II, GROUPS B, C, and D
Teledyne Analytical Instruments
Part I: i
Model 6600 Oil in Water Analyzer
Table of Contents
1 Introduction
1.1
1.2
1.3
1.4
Overview ........................................................................ 1-1
Typical Applications ....................................................... 1-1
Main Features of the Analyzer ....................................... 1-1
Operator Interface .......................................................... 1-2
1.4.1 UP/DOWN Switch ................................................ 1-3
1.4.2 ESCAPE/ENTER Switch ..................................... 1-3
1.5 Control Unit Interface Panel ........................................... 1-6
2 Installation
2.1 Unpacking the Control Unit ............................................ 2-1
2.2 Electrical Connections ................................................... 2-1
2.3 Testing the System ......................................................... 2-9
3 Operation
3.1 Introduction .................................................................... 3-1
3.2 Using the Controls ......................................................... 3-1
3.2.1 Mode/Function Selection ....................................... 3-2
3.3 The System Function ..................................................... 3-4
3.3.1 Setting up an Auto-Cal ........................................... 3-5
3.3.2 Password Protection .............................................. 3-6
3.3.2.1 Entering the Password ................................... 3-6
3.3.2.2 Installing or Changing the Password ............. 3-7
3.3.3 Logging Out ........................................................... 3-8
3.3.4 System Self-Diagnostic Test .................................. 3-9
3.3.5 The Model Screen ................................................. 3-10
3.3.6 Checking Linearity with Algorithm ......................... 3-10
3.3.7 Digital Flter Setup .................................................. 3-11
3.3.8 Homogenizer Function Setup ................................ 3-12
3.3.9 Hold/Track Setup ................................................... 3-13
3.3.10 Calibration/Hold Timer Setup ................................. 3-14
ii: Part I
Teledyne Analytical Instruments
Part I: Control Section
3.3.11 Analog 4 to 20 mA Output Calibration .................... 3-14
3.3.12 Manual Control of Filter & Solenoids ..................... 3-15
3.4 The Zero and Span Functions ....................................... 3-16
3.4.1 Zero Cal ................................................................. 3-17
3.4.1.1 Auto Mode Zeroing ........................................ 3-17
3.4.1.2 Manual Mode Zeroing .................................... 3-18
3.4.1.3 Zero Offset Calibration ................................... 3-19
3.4.1.4 Cell Failure .................................................... 3-19
3.4.2 Span Cal ................................................................ 3-21
3.4.2.1 Auto Mode Spanning ..................................... 3-21
4.4.2.2 Manual Mode Spanning ................................. 3-21
3.5 The Alarms Function ...................................................... 3-22
3.6 The Range Select Function ........................................... 3-24
3.6.1 Manual (Select/Define Range) Screen .................. 3-25
3.6.2 Auto Screen ........................................................... 3-25
3.6.3 Precautions ............................................................ 3-26
3.7 The Analyze Function .................................................... 3-27
3.8 Programming ................................................................. 3-28
3.8.1 The Set Range Screen .......................................... 3.29
3.8.2 The Curve Algorithm Screen ................................. 3-30
3.8.2.1 Manual Mode Linearization ........................... 3-30
3.8.2.2 Auto Mode Linearization ................................ 3-30
4 Maintenance
4.1 Fuse Replacement......................................................... 4-1
4.2 System Self Diagnostic Test ........................................... 4-2
4.3 Major Internal Components ............................................ 4-3
A Appendix
Model 6600 Specifications ..................................................... A-3
Teledyne Analytical Instruments
Part I: iii
Model 6600 Oil in Water Analyzer
iv: Part I
Teledyne Analytical Instruments
Oil in Water Analyzer
Part I: Control Section
Introduction
1.1
Overview
The Teledyne Analytical Instruments Model 6600 Control Section,
together with a 6600 Analysis Section, is versatile microprocessor-based
instrument.
Part I, of this manual covers the Model 6600 General Purpose, Bulkhead Mount Control Section. (The Analysis Section is covered in Part II of
this manual, and Oil in Water application is covered in Part III) The Control
Section is for indoor/outdoor use in hazardous environment only. The
Analysis Section (or Remote Section) can be designed for a variety of
hazardous environments. All Sections are mounted in a NEMA-4 enclosure
(24”x20”x10”).
1.2
Typical Applications when Configured
with the appropriate Sample System
A few typical applications of the Model 6600 are:
• Offshore platforms, Produced Water, Sea Water
• Waste Water, Tank Farms, Fuel Depots, Refinery Effluents
• Oil Chemical Separators
• On-Board Ship
• Boiler Return Steam Condensate
• Process Cooling Water
• Bilge/Deballast Water Treatment
• Water Soluble Oils
• All Aromatic Hydrocarbons
• Many Organic Hydrocarbons (Contact Factory)
Teledyne Analytical Instruments
Part I: 1-1
1 Introduction
1.3
Model 6600
Main Features of the Analyzer
The Model 6600 Photometric Analyzer is sophisticated yet simple to
use. The main features of the analyzer include:
•
A 2-line alphanumeric display screen, driven by microprocessor
electronics, that continuously prompts and informs the operator.
•
High resolution, accurate readings of concentration from low
ppm levels through to 100%. Large, bright, meter readout.
•
Versatile analysis over a wide range of applications.
•
Microprocessor based electronics: 8-bit CMOS microprocessor
with 32 kB RAM and 128 kB ROM.
•
Three user definable output ranges (from 0-1 ppm through
0-100 %) allow best match to users process and equipment.
•
Calibration range for convenient zeroing or spanning.
•
Auto Ranging allows analyzer to automatically select the proper
preset range for a given measurement. Manual override allows
the user to lock onto a specific range of interest.
•
Two adjustable concentration alarms and a system failure alarm.
•
Extensive self-diagnostic testing, at startup and on demand, with
continuous power-supply monitoring.
•
RS-232 serial digital port for use with a computer or other digital
communication device.
•
Analog outputs for concentration and range identification.
(0-1 V dc standard, and isolated 4–20 mA dc)
•
Superior accuracy.
•
Internal calibration-Manual or Automatic (optional).
1.4
Operator Interface
All controls and displays on the standard 6600 are accessible from
outside the housing. The instrument has two simple operator controls. The
operator has constant feedback from the instrument through an alphanumeric
display, and a digital LED meter. The displays and controls are described
briefly here and in greater detail in chapter 3. See Figure 1-1.
1-2: Part I
Teledyne Analytical Instruments
Oil in Water Analyzer
Part I: Control Section
1.4.1 UP/DOWN Switch
The UP/DOWN switch is used to select between any subfunctions displayed
on the VFD screen such as in the main menue, the system menue, the Alarm
menue, etc. When modifiable values are displayed on the VFD, the UP/DOWN
switch can be used to increment or decrement the values.
1.4.2 ESCAPE/ENTER Switch
The ESCAPE/ENTER switch is used to input the data, to enter a function,
or to exit a function displayed in the alphanumeric display:
•
Escape
Moves VFD display back to the previous screen in a
series. If none remains, returns to Analyze mode screen.
•
Enter
Within a menue: the funtion selected is entered moving on
to the next screen in a series.
With Value selected: Enters the value into the analyzer as
data. Advances cursor on VFD to the next operation.
In the Analyze mode: it calls the main menue. Functions
called out by the main menue:
-System
This function is a menu that calls a number
of functions that regulate the analyzer
operation.
-Span
This function spans the instrument.
-Zero
This function zeros the instrument.
-Alarms
This functions sets the alarm preferences.
-Range
This function selects whether analyzer is
autoranging or locked on one range.
-Standby Places the analyzer in a sleep mode.
WARNING:
The power cable must be disconnected to fully remove power
from the instrument.
Teledyne Analytical Instruments
Part I: 1-3
1 Introduction
Model 6600
Figure 1-1: Model 6600 Controls, Indicators, and Connectors
.
Digital Meter Display: The meter display is a Light Emitting Diode
LED device that produces large, bright, 7-segment numbers that are legible
in any lighting. It is accurate across all analysis ranges. The 6600 models
produce continuous readout from 0-10,000 ppm and then switch to
continuous percent readout from 1-100 %.
1-4: Part I
Teledyne Analytical Instruments
Oil in Water Analyzer
Part I: Control Section
Figure 1-2: Model 6600 Interface Panel
Teledyne Analytical Instruments
Part I: 1-5
1 Introduction
Model 6600
Alphanumeric Interface Screen: The backlit VFD screen is an easyto-use interface between operator and analyzer. It displays values, options,
and messages for immediate feedback to the operator.
1.5
Control Section Interface Panel
The Control Section interface panel, shown in Figure 1-2, contains the
electrical terminal blocks for external inputs and outputs. The input/output
functions are described briefly here and in detail in the Installation chapter of
this manual.
•
Power Connection
AC power source, 115VAC, 50/60 Hz
•
Analog Outputs
0-1 V dc concentration and 0-1 V dc
range ID. Isolated 4-20 mA dc and 4-20
mA dc range ID.
•
Alarm Connections
2 concentration alarms and 1 system
alarm.
•
RS-232 Port
Serial digital concentration signal output
and control input.
•
Remote Bench
Provides all electrical interconnect to the
Analysis Section.
Remote Span/Zero
Digital inputs allow external control of
analyzer calibration.
•
Calibration Contact
To notify external equipment that
instrument is being calibrated and
readings are not monitoring sample.
•
Range ID Contacts
Four separate, dedicated, range relay
contacts.
•
Network I/O
Serial digital communications for local
network access. For future expansion.
Not implemented at this printing.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the
Model 6600 is accurate to 2-3 %. Accordingly, internally scheduled calibrations can vary 2-3 % per day.
1-6: Part I
Teledyne Analytical Instruments
Oil in Water Analyzer
Part I: Control Section
Installation
Installation of Model 6600 Analyzers includes:
1. Unpacking, mounting, and interconnecting the Control/Analysis
Section
2. Making gas connections to the system
3. Making electrical connections to the system
4. Testing the system.
This chapter covers installation of the Control Section. (Installation of
the Analysis Section is covered in Part II of this manual.) The Oil in Water
application is covered in Part III.
2.1
Unpacking the Control/Analysis Unit
The analyzer is shipped with all the materials you need to install and
prepare the system for operation. Carefully unpack the Control/Analysis
Unit and inspect it for damage. Immediately report any damage to the shipping agent. Figure 2-2: Required Front Door Clearance
Allow clearance for the door to open in a 90-degree arc of radius 15.5
inches. See Figure 2-2.
Figure 2-2: Required Front Door
Clearance
”
.5
15
2.2
Electrical Connections
Figure 2-3 shows the Control/Analysis Unit interface panel. Connections for power, communications, and both digital and analog signal outputs
are described in the following paragraphs. Wire size and maximum length
data appear in the Drawings at the back of this manual.
Teledyne Analytical Instruments
Part I: 2-1
2 Installation
Model 6600
Figure 2-3: Interface Panel of the Model 6600 Control Section
For safe connections, ensure that no uninsulated wire extends
outside of the terminal blocks. Stripped wire ends must insert completely
into terminal blocks. No uninsulated wiring should come in contact with
fingers, tools or clothing during normal operation.
Primary Input Power: The power supply in the Model 6600 will
accept a 115 Vac, 50/60 Hz power source. See Figure 2-4 for detailed
connections.
DANGER: Power is applied to the instrument's circuitry as
long as the instrument is connected to the power
source. The standby function switches power on or
off to the displays and outputs only.
115VAC
Figure 2-4: Primary Input Power Connections
2-2: Part I
Teledyne Analytical Instruments
Oil in Water Analyzer
Part I: Control Section
Fuse Installation: The fuse holders accept 5 x 20 mm, 4.0 A, T
type (slow blow) fuses. Fuses are not installed at the factory. Be sure to
install the proper fuse as part of installation (See Fuse Replacement in
chapter 4, maintenance.)
Analog Outputs: There are eight DC output signal connectors on
the ANALOG OUTPUTS terminal block. There are two connectors per
output with the polarity noted. See Figure 2-5.
The outputs are:
0–1 V dc % of Range: Voltage rises linearly with increasing sample concentration, from 0 V at 0% to 1 V at 100%. (Full
scale = 100% programmed range.)
0–1 V dc Range ID:
0.25 V = Range 1, 0.5 V = Range 2, 0.75 V =
Range 3.
4–20 mA dc % Range: (-M Option) Current increases linearly with increasing sample concentration, from 4 mA at 0% to 20
mA at full scale 100%. (Full scale = 100% of
programmed range.)
4–20 mA dc Range ID: (-M Option) 8 mA = Range 1, 12 mA = Range 2,
16 mA = Range 3.
Figure 2-5: Analog Output Connections
Examples:
Teledyne Analytical Instruments
Part I: 2-3
2 Installation
Model 6600
The analog output signal has a voltage which depends on the sample
concentration AND the currently activated analysis range. To relate the
signal output to the actual concentration, it is necessary to know what range
the instrument is currently on, especially when the analyzer is in the
autoranging mode.
The signaloutput for concentration is linear over currently selected
analysis range. For example, if the analyzer is set on a range that was
defined as 0-10 %, then the output would be as shown in Table 2-1.
Table 2-1: Analog Concentration Output-Examples
Concentration
%
Voltage Signal
Output (V dc)
Current Signal
Output (mA dc)
0
0.0
4.0
1
0.1
5.6
2
0.2
7.2
3
0.3
8.8
4
0.4
10.4
5
0.5
12.0
6
0.6
13.6
7
0.7
15.2
8
0.8
16.8
9
0.9
18.4
10
1.0
20.0
To provide an indication of the range, a second pair of analog output
terminals are used. They generate a steady preset voltage (or current when
using the current outputs) to represent a particular range. Table 2-2 gives the
range ID output for each analysis range.
2-4: Part I
Teledyne Analytical Instruments
Oil in Water Analyzer
Part I: Control Section
Table 2-2: Analog Range ID Output - Example
Range
Voltage (V)
Current (mA)
Range 1
0.25
8
Range 2
0.50
12
Range 3
0.75
16
Alarm Relays:
There are three alarm-circuit connectors on the alarm relays block
(under RELAY OUTPUTS) for making connections to internal alarm relay
contacts. Each provides a set of Form C contacts for each type of alarm.
Each has both normally open and normally closed contact connections. The
contact connections are indicated by diagrams on the rear panel. They are
capable of switching up to 3 ampers at 250 V AC into a resistive load
(Figure 2-6).
Normally closed
Normally open
Moving contact
Normally open
Moving contact
Figure 2-6: Types of Relay Contacts
The connectors are:
Threshold Alarm 1:
• Can be configured as high (actuates when
concentration is above threshold), or low
(actuates when concentration is below thresh old).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
Teledyne Analytical Instruments
Part I: 2-5
2 Installation
Model 6600
Threshold Alarm 2:
System Alarm:
• Can be configured as high (actuates when concentration is above threshold), or low (actuates when
concentration is below threshold).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
Actuates when DC power supplied to circuits is
unacceptable in one or more parameters. Permanently
configured as fail-safe and latching. Cannot be defeated. Actuates if self test fails.
To reset a System Alarm during installation, disconnect power to the instrument and then reconnect it
Further detail can be found in chapter 3, section 3-5.
Digital Remote Cal Inputs
Remote Zero and Span Inputs: The REMOTE SPAN and REMOTE ZERO inputs are on the DIGITAL INPUT terminal block. They
accept 0 V (OFF) or 24 V dc (ON) for remote control of calibration (See
Remote Calibration Protocol below.)
Zero:
Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the ZERO mode. Either side may be
grounded at the source of the signal. 0 to 1 volt across the
terminals allows ZERO mode to terminate when done. A
synchronous signal must open and close the external zero
valve appropriately. See Remote Probe Connector at end of
section 3.3. (With the -C option, the internal valves automatically operate synchronously).
Span:
Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the SPAN mode. Either side may be
grounded at the source of the signal. 0 to 1 volt across the
terminals allows SPAN mode to terminate when done. A
synchronous signal must open and close the external span
valve appropriately. See Remote Probe Connector at end of
section 3.3. (With the -C option, the internal valves automatically operate synchronously.)
Cal Contact: This relay contact is closed while analyzer is spanning
and/or zeroing. (See Remote Calibration Protocol below.)
2-6: Part I
Teledyne Analytical Instruments
Oil in Water Analyzer
Part I: Control Section
Remote Calibration Protocol: To properly time the Digital Remote
Cal Inputs to the Model 6600 Analyzer, the customer's controller must
monitor the Cal Relay Contact.
When the contact is OPEN, the analyzer is analyzing, the Remote Cal
Inputs are being polled, and a zero or span command can be sent.
When the contact is CLOSED, the analyzer is already calibrating. It
will ignore your request to calibrate, and it will not remember that request.
Once a zero or span command is sent, and acknowledged (contact
closes), release it. If the command is continued until after the zero or span is
complete, the calibration will repeat and the Cal Relay Contact (CRC) will
close again.
For example:
1) Test the CRC. When the CRC is open, Send a zero command
until the CRC closes (The CRC will quickly close.)
2) When the CRC closes, remove the zero command.
3) When CRC opens again, send a span command until the CRC
closes. (The CRC will quickly close.)
4) When the CRC closes, remove the span command.
When CRC opens again, zero and span are done, and the sample is
being analyzed.
Note: The Remote Bench terminal strip (section 3.6 Part III) provides
signals to ensure that the zero and span gas valves will be
controlled synchronously.
Range ID Relays: Four dedicated RANGE ID CONTACT relays .
The first four ranges are assigned to relays in ascending order—Range 1 is
assigned to RANGE 1 ID, Range 2 is assigned to RANGE 2 ID, Range 3
is assigned to RANGE 3 ID, and Range 4 is assigned to RANGE 4 ID.
Network I/O: A serial digital input/output for local network protocol.
At this printing, this port is not yet functional. It is to be used in future
versions of the instrument.
RS-232 Port: The digital signal output is a standard RS-232 serial
communications port used to connect the analyzer to a computer, terminal, or
other digital device. The pinouts are listed in Table 2-3.
Table 2-3: RS-232 Signals
RS-232 Sig
DCD
RS-232 Pin Purpose
1
Data Carrier Detect
Teledyne Analytical Instruments
Part I: 2-7
2 Installation
Model 6600
RD
2
Received Data
TD
3
Transmitted Data
DTR
4
Data Terminal Ready
COM
5
Common
DSR
6
Data Set Ready
RTS
7
Request to Send
CTS
8
Clear to Send
RI
9
Ring Indicator
The data sent is status information, in digital form, updated every two
seconds. Status is reported in the following order:
• The concentration in percent
• The range is use (HI< MED< LO)
• The span of the range 0-100%, etc)
• Which alarm - if any - are disabled (AL-x DISABLED)
• Which alarms - if any - are tripped (AL-x ON)
Each status output is followed by a carriage return and line feed.
Three input functions using RS-232 have been implemented to date.
They are described in Table 2-4.
Table 2-4: Commands via RS-232 Input
Command
Description
as
Immediately starts an autospan.
az
Immediately starts an autozero.
st
Toggling input. Stops/Starts any status message output
from the RS-232, Until st is sent again.
The RS-232 protocol allows some flexibility in its implementation.
Table 2-5 lists certain RS-232 values that are required by the 6000B/6600.
Table 2-5: Required RS-232 Options
Parameter
Baud
Byte
Parity
Stop Bits
Message Interval
2-8: Part I
Setting
2400
8 bits
none
1
2 seconds
Teledyne Analytical Instruments
Oil in Water Analyzer
Part I: Control Section
Remote Bench and Solenoid Valves: The 6600 is a single-chassis
instrument. However, the REMOTE BENCH and SOLENOID RETURN
connectors are provided on the interface PCB. The Remote Bench is wired
at the factory as well as any optional solenoid valves included in the system.
2.3
Testing the System
After The Control/Analysis Unit is both installed and interconnected,
and the system gas and electrical connections are complete, the system is
ready to test. Before plugging the unit into its power sources:
• Check the integrity and accuracy of the gas connections. Make
sure there are no leaks.
• Check the integrity and accuracy of all electrical connections.
Make sure there are no exposed conductors
•
•
Warning:
Check that sample pressure typically between 0 and 30 psig,
according to the requirements of your process.
Turn homogenizer power potentiometer fully counter-clockwise
(OFF), see section 3.3.8 for operation of homogenizer.
Do not operate the “ultrasonic homogenizer” in the
instrument for more than one (1) minute without a
liquid sample properly flowing through the homogenizer.
Power up the system, and test it by performing the following operation:
1. Repeat the Self-Diagnostic Test.
2. Zero the instrument.
3. Span the instrument.
For steps 2 and 3, refer to part II for gas calibration, and part III for Oil
in Water application.
Teledyne Analytical Instruments
Part I: 2-9
2 Installation
2-10: Part I
Model 6600
Teledyne Analytical Instruments
Oil in Water Analyzer
Operation 3
Operation
3.1
Introduction
Although the Model 6600 is usually programmed to your application at the
factory, it can be further configured at the operator level, or even, cautiously,
reprogrammed. Depending on the specifics of the application, this might include
all or a subset of the following procedures:
• Setting system parameters:
• Establish a security password, if desired, requiring Operator
to log in (secure in safe file for referrence).
• Establish and start an automatic calibration cycle, if desired.
•
•
Routine Operation:
• Calibrate the instrument.
• Choose autoranging or select a fixed range of analysis.
• Set alarm setpoints, and modes of alarm operation (latching,
fail-safe, etc).
Program/Reprogram the analyzer:
• Define new applications.
• Linearize your ranges.
If you choose not to use password protection, the default password is
automatically displayed on the password screen when you start up, and you
simply press Enter for access to all functions of the analyzer.
3.2
Using the Controls
To get the proper response from these controls, turn the control toward the
desired action (ESCAPE or ENTER—DOWN or UP), and then release it.
Turn-and-release once for each action. For example, turn-and-release twice
toward UP to move the VFD screen two selections upwards on the list of
options (menu).
Teledyne Analytical Instruments
Part I
3-1
3 Operation
Model 6600
The item that is blinking on the screen is the item that is currently selectable
by choosing ENTER (turn-and-release toward ENTER with the ESCAPE/
ENTER control).
In these instructions, to ENTER means to turn-and-release toward ENTER, and To ESCAPE means to turn-and-release towards ESCAPE. To scroll
UP (or scroll DOWN) means to turn-and-release toward UP (or DOWN) as
many times as necessary to reach the required menu item.
3.2.1
Mode/Function Selection
After the instrument has been powered up, and its initilization routine
performed, the instrument will settle in the Analyze mode. To call up the Main
menu from the Analyze mode, toggle the Enter switch. To return to the Analyze
mode, toggle the Escape switch. The Main menu screens looks as shown
below:
SYSTEM SPAN ZERO
ALARM RANGE STBY
The Main menue screen is the top level in a series of screens used to
configure the analyzer. The DOWN/UP selects the different options displayed
in the VFD screen. The selectable option blinks on the VFD screen when you
reach the desired option, toggle the Enter switch.
The Escape switch takes you back up to hierarchy of screens until you
return back to the Analyze screen mode. Here is a brief description of the Main
Menu:
• System. The system function consists of nine subfunctions.
Four of these are for ordinary setup and operation:
• Setup an Auto-Cal
• Assign Passwords
• Log out to secure system
• Initiate a Self-Test
Three of the subfunctions do auxiliary tasks:
• Checking model and software version
• Adjust electronic filter of the signal
• Display more subfunctions
Two of these are for programming/reprogramming the analyzer:
• Define gas applications and ranges (Refer to programming
section, or contact factory.)
3-2
Part I
Teledyne Analytical Instruments
Oil in Water Analyzer
Operation 3
System
Dig_filt
Set Digital
Filter
SELF-TEST
Self-Test in
Progress
Self-Test
Results
PWD
Enter
Password
Change
Yes/No
LOGOUT
Secure System
setup not
allowed
Change
Password
Verify
Password
Enter
MORE
AUTOCAL
Span/Zero status
and
<>setup
HMGNZR
Set Ultrasonic
Homogenizer
ON/OFF
TRACK or
HOLD
Set track or
hold output
CAL-HOLD
TIMER
Set cal. hold
and sample
hold timer
Span/Zero timing
and on/off
Enter
Enter
Enter
MORE
Verify data
Points
ALGORITHM
Select range
Display gas use
and range
Define
Application/
Range
APPLICATION
Select range
MODEL
Display
Model/Version
Enter
OUTPUT_CAL
Zero Analog
Output
Enter
Select
Verify/Setup
Enter
Enter
Input/Output
Enter
Enter Span
gas value
Enter
Auto/
Manual
linear Cal.
Enter
Calibrate
Analog
Output
Enter
Figure 3-1: Hierarchy of System Functions and Subfunctions
Teledyne Analytical Instruments
Part I
3-3
3 Operation
Model 6600
•
•
•
•
•
Use the Curve Algorithm to linearize output. (Refer to
programming section, or contact factory.)
Zero. Used to set up a zero calibration.
Span. Used to set up a span calibration.
Alarms. Used to set the alarm setpoints and determine whether
each alarm will be active or defeated, HI or LO acting, latching,
and/or fail-safe.
Range. Used to set up four analysis ranges that can be switched
automatically with autoranging or used as individual fixed
ranges.
Any function can be selected at any time in the analyze mode (unless
password restrictions apply). The order as presented in this manual is
appropriate for an initial setup.
Each of these functions is described in greater detail in the following procedures. The VFD screen text that accompanies each operation is reproduced, at
the appropriate point in the procedure, in a Monospaced type style.
3.3
The System Function
The subfunctions of the System function are described below. Specific
procedures for their use follow the descriptions:
• Dig_Filt: Adjust how much digital filtering should be on the
signal
• SELF-TEST: Performs a self-diagnostic test to check the integrity
of the power supplies, outputs, detector signal and preamplifier.
•
•
•
•
•
•
•
3-4
PWD: Login security system for accessing to the setup functions.
LOGOUT: Prevents an unauthorized tampering with analyzer
settings.
AUTOCAL: Set the automatic calibrated timer schedule for Zero
and Span cycling.
HMGNZR: Turn Ultrasonic homogenizer ON and OFF on the
analyze mode.
TRACK: Set the system reading to be held or followed by the
concentration “gas or filter” during calibration.
CAL-HOLD-TIMER: Set the timing for calibration holding and
timing for the sample reading after return to analyze mode.
ALGORITHM: Linearize the output for nonlinear characteristic.
Part I
Teledyne Analytical Instruments
Oil in Water Analyzer
•
•
•
Operation 3
APPLICATION: Used to define the analysis ranges and application
(gas used).
MODEL: Displays model number and software version.
OUTPUT_CAL: 4-20 MA: Adjust 4 and 20 mA output.
The hierarchy of the system menu is shown in figure 3-1.
3.3.1 Setting up an AUTO-CAL
When proper automatic valving is connected, the Analyzer can cycle itself
through a sequence of steps that automatically zero and span the instrument.
Note: Before setting up an AUTO-CAL, be sure you understand the
Zero and Span functions as described in section 4.4, and
follow the precautions given there.
Note: If you require highly accurate AUTO-CAL timing, use external
AUTO-CAL control where possible. The internal clock in the
Model 6600 is accurate to 2-3 %. Accordingly, internally scheduled calibrations can vary 2-3 % per day.
To setup an Auto–Cal cycle:
Choose System from the main manu. TheVFD will display five subfunctions.
DIG_FILT SELF—TEST
PWD LOGOUT MORE
Select MORE and Enter
AUTOCAL HMGNZR
CAL-HOLD-TIMER
HOLD
MORE
Use UP/DOWN to blink AUTO—CAL, and Enter. A new screen for
ZERO/SPAN set appears.
ZERO in
SPAN in
Ød
Ød
Øh off
Øh off
Use UP/DOWN to blink ZERO (or SPAN), then Enter again. (You won’t
be able to set OFF to ON if a zero interval is entered.) A Span Every ...
(or Zero Every ...) screen appears.
Zero schedule: OFF
Day:
Ød Hour:
Øh
Teledyne Analytical Instruments
Part I
3-5
3 Operation
Model 6600
Use UP/DOWN to set the day interval, hour interval, then Enter
Enter to turn ON the SPAN and/or ZERO cycles (to activate AUTO–CAL).
Use the UP/DOWN to toggle the field between ON and OFF. Press Enter to
return to The AUTO-CAL menu. You should be able to see that the screen has
been updated with your new input. Escape to return to the System menu.
For Oil & Water Samples only setting of the Zero schedule is needed. Instrument will automatically perform a Span at the end of a scheduled Zero.
If instrument is turned off, the next time the instrument is powered, the
instrument will automatically perform a calibration cycle after 3 minutes of
entering the sample mode if AUTOCAL functions were on prior to shut down.
3.3.2 Password Protection
Before a unique password is assigned, the system assigns TAI by default.
This password will be displayed automatically. The operator just uses the Enter
switch to be allowed total access to the instrument’s features.
If a password is assigned, then setting the following system parameters can
be done only after the password is entered: alarm setpoints, assigning a new
password, range/application selections, and curve algorithm linearization.
(APPLICATION and ALGORITHM are covered in the programming section.)
However, the instrument can still be used for analysis or for initiating a self-test
without entering the password. To defeat security the password must be
changed back to TAI.
NOTE: If you use password security, it is advisable to keep a copy of
the password in a separate, safe location.
3.3.2.1
Entering the Password
To install a new password or change a previously installed password, you
must key in and ENTER the old password first. If the default password is in
effect, using ENTER three times will enter the default TAI password for you.
Enter the System menu...
DIG_FILT AUTO—CAL
PWD LOGOUT MORE
Use theUP/DOWN to scroll the blinking over to PWD, and Enter to select
the password function. Either the default TAI password or AAA place holders for
an existing password will appear on screen depending on whether or not a
password has been previously installed.
3-6
Part I
Teledyne Analytical Instruments
Oil in Water Analyzer
Operation 3
Enter password:
T A I
or
Enter password:
A A A
The screen prompts you to enter the current password. If you are not using
password protection, Enter three times to accept TAI as the default password.
If a password has been previously installed, enter the password using the UP/
DOWN SWITCH to change the first letter.
Use Enter to move to the next letter. You cannot go back. If a mistake is
made, Escape to the System menu and return.
When you finish adjusting the last letter, toggle the Enter switch
In a few seconds, you will be given the opportunity to change this password or keep it and go on.
Change Password?
=Yes
=No
Escape to move on, or proceed as in Changing the Password, below.
3.3.2.2
Installing or Changing the Password
If you want to install a password, or change an existing password, proceed
as above in Entering the Password. When you are given the opportunity to
change the password:
Change Password?
=Yes
=No
nter to change the password (either the default TAI or the previously
assigned password), or Escape to keep the existing password and move on.
If you chose Enter to change the password, the password assignment
screen appears.
Select new password
T A I
Enter the password using the UP/DOWN SWITCH to change the letters to
the new password. The full set of 94 characters available for password use are
shown in the table below.
Characters Available for Password Definition:
A
K
U
B
L
V
C
M
W
D
N
X
E
O
Y
F
P
Z
G
Q
[
Teledyne Analytical Instruments
H
R
¥
I
S
]
J
T
^
Part I
3-7
3 Operation
_
i
s
}
)
3
=
Model 6600
`
j
t
→
*
4
>
a
k
u
!
+
5
?
b
l
v
"
'
6
@
c
m
w
#
7
d
n
x
$
.
8
e
o
y
%
/
9
f
p
z
&
0
:
g
q
{
'
1
;
h
r
|
(
2
<
When you have finished typing the new password, press Enter. A verification screen appears. The screen will prompt you to retype your password for
verification.
Enter PWD To Verify:
A A A
Use the UP/DOWN to retype your password and Enter at the end of
each letter. Your password will be stored in the microprocessor and the system
will immediately switch to the Analyze screen, and you now have access to all
instrument functions.
If all alarms are defeated, the Analyze screen appears as:
1.95
nR1:
ppm SO2
Ø — 1Ø Anlz
If an alarm is tripped, the second line will change to show which alarm it is:
1.95
AL—1
ppm SO2
NOTE:If you log off the system using the LOGOUT function in the
system menu, you will now be required to re-enter the password to gain access to Alarm, and Range functions.
3.3.3 Logging Out
The LOGOUT function provides a convenient means of leaving the analyzer in a password protected mode without having to shut the instrument off. By
entering LOGOUT, you effectively log off the instrument leaving the system
protected against use until the password is reentered. To log out, enter the
System menu.
DIG_FILT SELF-TEST
PWD LOGOUT MORE
Use theUP/DOWN to position the blinking over the LOGOUT function,
and Enter to Log out. The screen will display the message:
3-8
Part I
Teledyne Analytical Instruments
Oil in Water Analyzer
Operation 3
Protected until
password entered
After two seconds it will return to the System menu.
3.3.4 System Self-Diagnostic Test
The Model 6600 has a built-in self-diagnostic testing routine. Pre-programmed signals are sent through the power supply, output board, preamp
board and sensor circuit. The return signal is analyzed, and at the end of the test
the status of each function is displayed on the screen, either as OK or as a
number between 1 and 1024. (See System Self Diagnostic Test in chapter 5
for number code.)
Note: The sensor will always show failed unless Zero fluid is present
in the sampling cell at the time of the SELF-TEST input thru the
sample inlet.
The self diagnostics are run automatically by the analyzer whenever the
instrument is turned on, but the test can also be run by the operator at will. If
any of the functions fails, the System Alarm is tripped, but is not tripped at the
startup because it might give false alarm after a power failure. To initiate a self
diagnostic test during operation enter the System menu.
DIG_FILT SELF-TEST
PWD LOGOUT MORE
Use the UP/DOWN again to move the blinking to the SELF–TEST and
Enter. The screen will follow the running of the diagnostic.
RUNNING DIAGNOSTIC
Testing Preamp — Cell
When the testing is complete, the results are displayed.
Power: OK Analog: OK
Cell:
2 Preamp: 3
The module is functioning properly if it is followed by OK. A number
indicates a problem in a specific area of the instrument. Refer to Chapter 5
Maintenance and Troubleshooting for number-code information. The results
screen alternates for a time with:
Press Any Key
To Continue...
Then the analyzer returns to the initial System screen.
Teledyne Analytical Instruments
Part I
3-9
3 Operation
Model 6600
3.3.5 The Model Screen
Enter the System menu, select more and Enter. The second screen appears. Select more again and Enter. In the third screen select MODEL. With
MODEL blinking, Enter. The screen displays the manufacturer, model, and
software version information. Escape to return to the System menu.
3.3.6 Checking Linearity with ALGORITHM
From the System Function screen, select ALGORITHM, and Enter.
sel rng to set algo:
—> Ø1 Ø2 Ø3
<—
Use the UP/DOWN switch to select the range: 01, 02, or 03. Then
Enter. (Some ranges may not be available, depending on your application, but
at least range 01 should be).
Gas Use:
SO2
Range:
Ø — 10%
Enter again.
Algorithm setup:
VERIFY
SET UP
Select and Enter VERIFY to check whether the linearization has been
accomplished satisfactorily.
Dpt
Ø
INPUT
Ø.ØØ
OUTPUT
Ø.ØØ
The leftmost digit (under Dpt) is the number of the data point being monitored. Use the UP/DOWN switch to select the successive points.
The INPUT value is the input to the linearizer. It is the simulated output of
the analyzer. You do not need to actually flow gas.
The OUTPUT value is the output of the linearizer. It should be the ACTUAL
concentration of the span gas being simulated.
If the OUTPUT value shown is not correct, the linearization must be corrected. Press ESCAPE to return to the previous screen. Select and Enter SET UP
to Calibration Mode screen. (set-up will not work without a PC being connected
to the analyzer)
Select algorithm
mode : AUTO
3-10
Part I
Teledyne Analytical Instruments
Oil in Water Analyzer
Operation 3
There are two ways to linearize: AUTO and MANUAL: The auto mode
requires as many calibration gases as there will be correction points along the
curve. The user decides on the number of points, based on the precision required.
The manual mode only requires entering the values for each correction
point into the microprocessor via the front panel buttons. Again, the number of
points required is determined by the user.
NOTE: If input and output are set to 0.00 for all data points, it
might be that your application is linear.
3.3.7 Digital Filter Setup
The 6600 has the option of decreasing or increasing the amount filtering on
the signal. This feature enhances the basic filtering done by the analog circuits by
setting the amount of digital filtering effected by the microprocessing. To access
the digital filter setup, you must:
1.
Enter the System menu
DIG_FILT
PWD LOGOUT
2.
SELF-TEST
MORE
DIG_FILT will flash, ENTER,
Weight of digital
Filter:
9
3.
The number on the second row will flash and can be set by
using the Up/Down switch
4.
Press Escape to return to the System menu.
The settings go from zero, no digital filtering, to 10, maximum digital filtering. The default setting is 8 and that should suffice for most applications. In
some applications where speeding the response time with some trade off in noise
is of value, the operator could decrease the number of the digital filter. In
applications where the signal is noisy, the operator could switch to a higher
number; the response time is slowed down though.
90% response time on the different settings to a step input is shown below.
This response time does not include the contributions of the bench sampling
system and the preamplifier near the detector.
Setting
90% Response time
(seconds)
0
4.5
Teledyne Analytical Instruments
Part I
3-11
3 Operation
Model 6600
1
4.5
2
5.0
3
5.0
4
5.5
5
7.0
6
9.0
7
14.0
8
25.0
9
46.0
10
90.0
At a setting of “zero”, the response time is purely set by the electronics to
4.5 seconds. The numbers above can and will change depending on application
and they merely serve to illustrate the effect of the digital filter.
3.3.8 Homogenizer Function Setup
Depending on the application, the 6020 sampling system may have an
Ultra Sonic Homogenizer. The function of this part is to prevent the oil in the
water from clumping together. The homogenizer should turn on after the initial
warm up and self-diagnostic period when the analyzer enters the Analyze mode.
The homogenizer will turn off automatically as soon as the analyzer enters the
Zero and Span mode, turning back on at the end.
Under some conditions, it might be desirable to manually turn the Ultra
Sonic Homogenizer off. The homogenizer set up function is provided in the
System menu. To access the Homogenizer function setup:
1. Enter the System menu.
2. Select MORE in the first System menu using the UP/DOWN switch.
3. Select HMGNZR in the second System menu, the screen will display
Set Ultra Sonic
Homogenizer:
ON
4. By using the UP/DOWN switch the homogenizer can be toggled on and off.
5. Enter or Escape to return to the analyze mode.
3-12
Part I
Teledyne Analytical Instruments
Oil in Water Analyzer
Operation 3
Every time the power is cycled, the homogenizer defaults to ON. So if
homogenizer was off and the power is cycled, the homogenizer will turn on.
Warning: Do not operate the “ultrasonic homogenizer” in the
instrument for more than one (1) minute without a
liquid sample properly flowing through the homogenizer.
3.3.9
Hold/Track Setup
The 6600 has ability to disable the analog outputs and freeze the display
while undergoing a scheduled or remote calibration. The 6600 will track
changes in the concentration if calibration is started through the front
panel. To setup this feature, the operator must:
1.
Enter the System menu:
DIG_FILT
SELF-TEST
PWD LOGOUT
MORE
2.
Using the UP/DOWN switch, select MORE and Enter. The
Second System screen appears:
AUTOCAL HMGNZR TRACK
CAL-HOLDER-TIMER
MORE
or
AUTOCAL HMGNZR HOLD
CAL-HOLD-TIMER
MORE
3.
The option on the right of the first row can be set to TRACK or
HOLD by toggling Enter switch. By selecting the TRACK option, the analog
outputs are enabled and with the display will track the concentration changes
while the instrument is undergoing scheduled or remote calibration (either zero or
span). By selecting the HOLD option, the analog outputs and display are
disabled and will not track the concentration changes while the instrument is
undergoing scheduled or remote calibration (either zero or span). In the HOLD
option, the analog outputs and display will freeze on the last reading before
entering calibration.
The analog outputs are both 0 to 1 volt outputs and both 4 to 20 mA
outputs.
Teledyne Analytical Instruments
Part I
3-13
3 Operation
Model 6600
3.3.10 Calibration/Hold Timer Setup
This Calibration Timer lets the operator adjust the time the instrument
purges the calibration gas prior to actually starting the calibration computations.
The Sample timer lets the operator adjust the time the instrument purges sample
gas after finishing a calibration before it lets the analog outputs and display track
the change in concentration.
This function and the TRACK/HOLD feature will prevent false alarms
while performing remote or autoscheduled calibrations. These functions are not
applicable if the calibration is initiated through the front panel. To enter the
Calibration/Hold Timer function, you must:
1.
Enter the System menu:
DIG_FILT
SELF-TEST
PWD LOGOUT
MORE
2.
Using theUP/DOWN switch, select MORE and press Enter:
The Second System screen appears:
AUTOCAL HMGNZR TRACK
CAL-HOLD-TIMER
MORE
or
AUTOCAL HMGNZR HOLD
CAL-HOLD-TIMER
MORE
3.
Select with the UP/DOWN switch CAL-HOLD-TIMER,
and press the Enter key to access this function menu:
Calbrt hold:
3 min
Sample hold:
1 min
The calibration hold time is set on the first row, while the sample hold
time is set on the second row. To select one or the other, use the Right or
Left keys. To modify the time of either timer, use the Up or Down keys.
The time is in the minutes.
3.3.11 Analog 4-20mA Output Calibartion
This function will let the operator calibrate the 4 to 20 mA analog output to
match the display reading. A DMM configure as a DC ammeter is needed. The
DMM should be connected across the output terminals of the 4 to 20 mA output
to monitor the output current. To enter the 4 to 20 mA output adjust function,
you must:
1.
3-14
Part I
Enter the System menu:
Teledyne Analytical Instruments
Oil in Water Analyzer
Operation 3
DIG_FILT
SELF-TEST
PWD LOGOUT MORE
2.
Using the Right or Left arrow keys, select MORE and press
Enter. The second System screen appears:
AUTOCAL HMGNZR TRACK
CAL-HOLD-TIMER
MORE
or
AUTOCAL HMGNZR HOLD
CAL-HOLD-TIMER
MORE
3.
Using the Right or the Left arrow keys, select MORE and
press Enter. The third System screen appears:
ALGORITHM APPLICATION
MODEL OUT_CAL
ANLZ
4.
Select OUTPUT_CAL and Enter
Use UP/DOWN arrow to
Adjust 4 ma:
250
The number on the second row is the setpoint of the 4 mA output. It is
analogous to a potentiometer wiper. The number can be set anywhere from 0
to 500. The default is 250, in the middle. At the default setting, the output
should be very close to 4 mA. If not, slowly adjust the number using the Up or
the Down keys until DMM reads 4.00 mA. Enter when done.
5.
A screen similar to the one above will appear and the DMM
should read close to 20 mA. If not, slowly adjust the number using the Up or
Down key until DMM reads 20.0 mA. Enter when done to return to system
menu.
The range of adjustment is approximately +/- 10% of scale (+/- 1.6 ma).
Since the 4 to 20 mA output is tied to the 0 to 1 volt output, this function can be
used to calibrate the 0 to 1 volt output, if the 4 to 20 mA output is not used. By
using a digital Volt meter on the 0-1 Volt output.
3.3.12 Manual control of filter and solenoids
For troubleshooting purposes, you have manual access to control
calibration filter and solenoid on the Analyze mode. To have manual access to
the calibration filter and solenoid:
Teledyne Analytical Instruments
Part I
3-15
3 Operation
Model 6600
-Enter the System Menu
-Select MORE on the first and second System menu screens.
-In the last System Menu screen you will see:
ALGORITHM APPLCATION
MODEL OUT_CAL ANLZ
-Select the last field “ANLZ” using the Up/Down switch.
-Press Enter to change the mode of the filter and the solenoid. The sequence is as follows:
1. ZERO: Sets the Filter and solenoid in the zero mode (span filter off,
zero solenoid on).
2. SPN1: Sets the Filter and solenoid in the span mode (span filter on,
zero solenoid on).
3. SPN2: Sets second span filter on (usually not installed) and zero
solenoid on
4. SPNB: Sets both span filter on (usually only one filter installed) and
zero solenoid on.
5. ANLZ: Returns Filters and solenoid to the Analyze mode (span filter
off, zero solenoid off).
-Press Escape to see the effect
NOTE: FOR PROPER OPERATION OF THE ANALYZER
RETURN TO ANLZ MODE.
3.4
The Zero and Span Functions
The Model 6600 can have as many as three analysis ranges plus a special
calibration range (Cal Range). Calibrating any one of the ranges will automatically calibrate the other ranges.
CAUTION: Always allow one hour warm-up time before calibrating, if your analyzer has been disconnected from its
power source. This does not apply if the analyzer
was plugged in but was in STANDBY.
The analyzer is calibrated using zero, and span gases.
3-16
Part I
Teledyne Analytical Instruments
Oil in Water Analyzer
Operation 3
Note: Shut off the gas pressure before connecting it to the analyzer,
and be sure to limit pressure to 40 psig or less when turning it
back on.
Readjust the gas pressure into the analyzer until the flowrate through the
Sample Cell settles between 50 to 500 cc/min (approximately 0.1 to 0.4
SCFH).
Note: Always keep the calibration gas flow as close to the flowrate of
the sample gas as possible
3.4.1 Zero Cal
The Zero function on the main menu is used to enter the zero calibration
function. Zero calibration can be performed in either the automatic or manual
mode.
Make sure the zero fluid is flowing to the instrument. If you get a CELL
CANNOT BE BALANCED message while zeroing skip to section 3.4.1.3.
3.4.1.1
Auto Mode Zeroing
Observe the precautions in sections 3.4 and 3.4.1, above. Enter the zero
function mode. The screen allows you to select whether the zero calibration is to
be performed automatically or manually. Use the UP/DOWN switch to toggle
between AUTO and MAN zero settling. Stop when AUTO appears, blinking, on
the display.
Select zero
mode: AUTO
Enter to begin zeroing.
####.## ppm OIL
Slope=#.###
C—Zero
The beginning zero level is shown in the upper left corner of the display. As
the zero reading settles, the screen displays and updates information on Slope=
in percent/second (unless the Slope starts within the acceptable zero range and
does not need to settle further). The system first does a coarse zero, shown in
the lower right corner of the screen as C—Zero, for 3 min, and then does a fine
zero, and displays F—Zero, for 3 min.
Then, and whenever Slope is less than 0.01 for at least 12 sec, instead of
Slope you will see a countdown: 9 Left, 8 Left, and so fourth. These are
Teledyne Analytical Instruments
Part I
3-17
3 Operation
Model 6600
software steps in the zeroing process that the system must complete, AFTER
settling, before it can go back to Analyze. Software zero is indicated by S–
Zero in the lower right corner.
NOTE: In a Oil/Water sampling system, when performing a scheduled zero, instrument will go to span mode automatically (when span flag
option has been purchased).
####.## ppm
4 Left=#.###
OIL
S—Zero
The zeroing process will automatically conclude when the output is within
the acceptable range for a good zero. Then the analyzer automatically returns to
the Analyze mode.
3.4.1.2
Manual Mode Zeroing
Enter the Zero function. The screen that appears allows you to select
between automatic or manual zero calibration. Use the UP/DOWN switch
between AUTO and MAN zero settling. Stop when MANUAL appears, blinking,
on the display.
Select zero
mode: MANUAL
Enter to begin the zero calibration. After a few seconds the first of three
zeroing screens appears. The number in the upper left hand corner is the firststage zero offset. The microprocessor samples the output at a predetermined
rate.
####.##
ppm OIL
Zero adj:2048 C—Zero
The analyzer goes through C–Zero, F–Zero, and S–Zero. During C–Zero
and F–Zero, use the UP/DOWN SWITCH to adjust displayed Zero adj: value
as close as possible to zero. Then,Enter.
S–Zero starts. During S–Zero, the Microcontroller takes control as in Auto
Mode Zeroing, above. It calculates the differences between successive samplings and displays the rate of change as Slope= a value in parts per million per
second (ppm/s).
####.##
ppm
OIL
Slope=#.###
S—Zero
Once zero settling completes, the information is stored in the analyzer’s
memory, and the instrument automatically returns to the Analyze mode.
3-18
Part I
Teledyne Analytical Instruments
Oil in Water Analyzer
Operation 3
3.4.1.3 Detector Failure
Detector failure in the 6600 is usually associated with inability to zero the
instrument with a reasonable voltage differential between the reference and
measure voltages. If this should ever happen, the 6600 system alarm trips, and
the LCD displays a failure message.
Detector cannot be balanced
Check your zero fluid
Before optical balancing:
a. Check your zero fluid to make sure it is within specifications.
b. Check for leaks downstream from the Sample Cell, where contamination may be leaking into the system.
c. Check flowmeter to ensure that the flow is no more than 200
SCCM for liquids and 1000CCM for gases.
d. Check temperature controller board.
e. Check sample temperature.
f. Check the Sample Cell for dirty windows.
g. Perform a Zero calibration in the manual mode.
h. Check for air bubbles in liquid applications.
If none of the above, proceed to perform an optical balance as described in
chapter 3, part II.
3.4.1.4 Zero Offset Calibration
To access this function, the instrument zero mode must be entered by
pushing the Zero key on the front panel of the control unit. The VFD display will
show the following menu selection:
Select zero
mode: AUTO
or
Select zero
mode: MAN
Select whether you want the instrument to do an automatic or manual zero. If
you do an automatic zero, the instrument does the zero by itself. If you do a
manual zero you must manually enter inputs to the instrument to accomplish the
zero, see in the corresponding section of the manual on how this two functions
differ.
When the Enter key is pressed, the following menu will appear:
Teledyne Analytical Instruments
Part I
3-19
3 Operation
Model 6600
Zero off: 0.0 ppm
to begin Zero
The offset value can be modified by using the Up/Down keys. Next section
shows how to select this value. Suffice to say that whatever value you enter, will
be automatically added to the reading. Thus, if you entered -0.1 ppm, at the end
of the zero the display will show -0.1 ppm.
Once the Enter key is pressed the instrument enters the zero mode. If you chose
AUTO zero mode, the instrument will do the work of bringing the reading back
to zero plus the offset value that was entered. If you chose MANual zero mode,
then you must enter input to the instrument as explained in the corresponding
section of the manual but with one difference: instead of bringing the display to
read zero, you must make the display read zero plus the value entered as offset.
How the offset value is selected:
To find out what the offset value should be, the intended zero calibration gas and
the a mix of the process background gas must be procured. This of course
assumes that the zero gas and the process background gas are very different and
that an offset will occur.
1. Let the intended zero calibration gas flow through the 6600 sample cell (this
assumes that you have started up you system as recommended by the manual or
technical personnel) and do a zero on the instrument. Leave the offset set to zero
value.
2. At the end of the zero function, make sure the analyser reads zero.
3. Flow the process background gas mix through the 6000 sample cell on the
Analyse mode. Wait for the reading to become stable. Write the reading down.
Change the sign of the reading: This is the offset to be entered.
4. Do a manual run to check. Reintroduce the zero calibration gas. Start a zero
on the analyser but this time enter the offset value.
5. At the end of the zero function, check that the instrument reads the entered
offset.
6. Reintroduce the process background gas mix to the 6000 sample cell in the
Analyse mode. It should read close to zero once the reading is stable (+/- 1%
error of full scale).
Spanning the 6600:
Since the instrument might be spanned with background gas the same as the
zero calibration gas, the span value to be entered should be the span
concentration plus the offset value (if the offset value has a minus sign then
algebraically it becomes a subtraction).
3-20
Part I
Teledyne Analytical Instruments
Oil in Water Analyzer
Operation 3
3.4.2 Span Cal
The Span function on the main menu is used to span calibrate the analyzer.
Span calibration can be performed in either the automatic or manual mode.
Make sure the span fluid is flowing to the instrument.
3.4.2.1
Auto Mode Spanning
Observe all precautions in sections 3.4 and 3.4.2, above. Enter the span
function. The screen that appears allows you to select whether the span calibration is to be performed automatically or manually. Use the UP/DOWN SWITCH
to toggle between AUTO and MAN span settling. Stop when AUTO appears,
blinking, on the display.
Select span
mode: AUTO
Enter to move to the next screen.
Span Val:
2Ø.ØØ %
twice to start
The unit field should be blinking first (%/ppm). Use the UP/DOWN switch
to set the proper unit of the span fluid and enter. Then, use the UP/DOWN
switch to set the concentration. When you have set the concentration of the
span fluid you are using, Enter to begin the Span calibration.
####.## ppm
Slope=#.###
OIL
Span
The beginning span value is shown in the upper left corner of the display.
As the span reading settles, the screen displays and updates information on
Slope. Spanning automatically ends when the span output corresponds, within
tolerance, to the value of the span gas concentration. Then the instrument automatically returns to the analyze mode.
3.4.2.2
Manual Mode Spanning
Enter the Span function. The screen that appears allows you to select
whether the span calibration is to be performed automatically or manually.
Select span
mode: MANUAL
Teledyne Analytical Instruments
Part I
3-21
3 Operation
Model 6600
Use the UP/DOWN switch to toggle between AUTO and MAN span
settling. Stop when MAN appears, blinking, on the display. Enter to move to
the next screen.
Span Val: 100 ppm
To begin span
The unit field should be blinking first (%/ppm). Use the UP/DOWN switch
to set the proper unit of the span fluid and enter. Then, use the UP/DOWN
switch to set the concentration.
When you have set the concentration of the span fluid you are using, Enter
to begin the Span calibration.
Once the span has begun, the microprocessor samples the output at a
predetermined rate. It calculates the difference between successive samplings
and displays this difference as Slope on the screen. It takes several seconds for
the first Slope value to display. Slope indicates rate of change of the Span
reading. It is a sensitive indicator of stability.
####.## ppm
Slope=#.###
OIL
Span
When the Span value displayed on the screen is sufficiently stable, Enter.
(Generally, when the Span reading changes by 1 % or less of the range being
calibrated for a period of 30 seconds it is sufficiently stable.) Once the span
ends, the calibration is stored in memory. The instrument then automatically
enters the Analyze function.
3.5
The Alarms Function
The Model 6600 is equipped with 2 fully adjustable set points concentration with two alarms and a system failure alarm relay. Each alarm relay has a set
of form “C" contacts rated for 3 amperes resistive load at 250 V ac. See Figure
in Chapter 2, Installation and/or the Interconnection Diagram included at the
back of this manual for relay terminal connections.
The system failure alarm has a fixed configuration described in chapter 2
Installation.
The concentration alarms can be configured from the front panel as either
high or low alarms by the operator. The alarm modes can be set as latching or
non-latching, and either failsafe or non-failsafe, or, they can be defeated
altogether. The setpoints for the alarms are also established using this function.
Decide how your alarms should be configured. The choice will depend
upon your process. Consider the following four points:
3-22
Part I
Teledyne Analytical Instruments
Oil in Water Analyzer
Operation 3
1. Which if any of the alarms are to be high alarms and which if any
are to be low alarms?
Setting an alarm as HIGH triggers the alarm when the
contaminant concentration rises above the setpoint. Setting an
alarm as LOW triggers the alarm when the contaminant
concentration falls below the setpoint.
Decide whether you want the alarms to be set as:
• Both high (high and high-high) alarms, or
• One high and one low alarm, or
• Both low (low and low-low) alarms.
2. Are either or both of the alarms to be configured as failsafe?
In failsafe mode, the alarm relay de-energizes in an alarm
condition. For non-failsafe operation, the relay is energized in an
alarm condition. You can set either or both of the concentration
alarms to operate in failsafe or non-failsafe mode.
3. Are either of the alarms to be latching?
In latching mode, once the alarm or alarms trigger, they will
remain in the alarm mode even if process conditions revert back
to non-alarm conditions. This mode requires an alarm to be
recognized before it can be reset. In the non-latching mode, the
alarm status will terminate when process conditions revert to nonalarm conditions.
4. Are either of the alarms to be defeated?
The defeat alarm mode is incorporated into the alarm circuit so
that maintenance can be performed under conditions which
would normally activate the alarms.
The defeat function can also be used to reset a latched alarm.
(See procedures, below.)
If you are using password protection, you will need to enter your password
to access the alarm functions. Follow the instructions in section 3.3.3 to enter
your password. Once you have clearance to proceed, enter the Alarm function.
Select the Alarm function on the main menu to enter the Alarm function.
Use UP/DOWN to select either AL1 or AL2. If you must change the %/ppm
units, keep using UP/DOWN until you get to the units to be modified. Use Enter
to toggle between % and ppm.
Teledyne Analytical Instruments
Part I
3-23
3 Operation
Model 6600
Use the UP/DOWN to choose the alarm again. Then Enter to move to
the next screen.
AL1: 1ØØØ
ppm HI
Dft:N Fs:N Ltch:N
Five parameters can be changed on this screen:
• Value of the alarm setpoint, AL1: ####
• Out-of-range direction, HI or LO
• Defeated? Dft:Y/N (Yes/No)
• Failsafe? Fs:Y/N (Yes/No)
• Latching? Ltch:Y/N (Yes/No).
• To define the setpoint use the UP/DOWN switch to change the
number. Holding down the key speeds up the incrementing or
decrementing. Enter when done, blinking cursor moves to the
next field.
• To set the other parameters use the UP/DOWN key, and then
Enter to move to the next field..
• Once the parameters for alarm 1 have been set, Enter the Alarms
function again, and repeat this procedure for alarm 2 (AL2).
• To reset a latched alarm, go to Dft– and then toggle either Up or
DOWN two times. (Toggle it to Y and then back to N.)
–OR –
Go to Ltch– and then toggle either UP two times or DOWN two
times. (Toggle it to N and back to Y.)
3.6
The Range Select Function
The Range function allows you to manually select the concentration range
of analysis (MANUAL), or to select automatic range switching (AUTO).
In the MANUAL screen, you are further allowed to define the high and low
(concentration) limits of each Range, and select a single, fixed range to run.
CAUTION: If this is a linearized application, the new range must
be within the limits previously programmed using the
System function, if linearization is to apply throughout the range. Furthermore, if the limits are too small
a part (approx 10 % or less) of the originally linearized range, the linearization will be compromised.
3-24
Part I
Teledyne Analytical Instruments
Oil in Water Analyzer
Operation 3
3.6.1 Manual (Select/Define Range) Screen
The Manual range-switching mode allows you to select a single, fixed
analysis range. It then allows you to redefine the upper and lower limits, for the
range.
Enter the Range function to start the Range function.
Select range
mode: MANUAL
If above screen displays, use the UP/DOWN switch to Select MANUAL,
and Enter.
Select range to run
—> Ø1 Ø2 Ø3 <—
NOTE: Oil in Water applications require single range (01) measurement.
Use the UP/DOWN switch to select the range: 01, 02, 03, or 04. Then
Enter.
Gas use:
OIL
Range:
Ø — 100 ppm
The high-end of the range field should blink first. Use UP/DOWN switch
to change the value of the field, Enter to move to the low-end of the range field.
Escape to return to the previous screen to select or define another range.
Enter to return the to the Analyze function.
3.6.2 Auto Screen
Autoranging will automatically set to the application that has at least two
ranges setup with the same gases.
In the autoranging mode, the microprocessor automatically responds to
concentration changes by switching ranges for optimum readout sensitivity. If the
upper limit of the operating range is reached, the instrument automatically shifts
to the next higher range. If the concentration falls to below 85% of full scale of
the next lower range, the instrument switches to the lower range. A corresponding shift in the DC concentration output, and in the range ID outputs, will be
noticed.
The autoranging feature can be overridden so that analog output stays on a
fixed range regardless of the contaminant concentration detected. If the conce-
Teledyne Analytical Instruments
Part I
3-25
3 Operation
Model 6600
tration exceeds the upper limit of the range, the DC output will saturate at 1 V dc
(20 mA at the current output).
However, the digital readout and the RS-232 output of the concentration
are unaffected by the fixed range. They continue to read beyond the full-scale
setting until amplifier saturation is reached. Below amplifier saturation, the
overrange readings are accurate UNLESS the application uses linearization over
the selected range.
The concentration ranges can be redefined using the Range function
Manual screen, and the application gases can be redefined using the System
function, if they are not already defined as necessary.
CAUTION: Redefining applications or ranges might require
relinearization and/or recalibration.
To setup automatic ranging:
Enter the Range function to start the Range function.
Select range
mode : AUTO
If above screen displays MAN, use the UP/DOWN switch to Select
AUTO, and Enter.
Press Escape to return to the previous Analyze Function.
3.6.3 Precautions
The Model 6600 allows a great deal of flexibility in choosing ranges for
automatic range switching. However, there are some pitfalls that are to be
avoided.
Ranges that work well together are:
• Ranges that have the same lower limits but upper limits that differ
by approximately an order of magnitude
• Ranges whose upper limits coincide with the lower limits of the
next higher range
• Ranges where there is a gap between the upper limit of the range
and the lower limit of the next higher range.
Range schemes that are to be avoided include:
• Ranges that overlap
3-26
Part I
Teledyne Analytical Instruments
Oil in Water Analyzer
Operation 3
•
Ranges whose limits are entirely within the span of an adjoining
range.
• Ranges where the zero is suppressed, is 1-10, 1-100, etc,
however, 80-100, 90-100 is ok where the zero gas is actually
100% concentration and the calibration is inverted.
• In Oil and Water applications, because the range and cell path are
pertinent to the water background and preparation of zero fluid
by the sample system, the autorange feature should not be used.
Only single range is recommended.
Figure 3-2 illustrates these schemes graphically.
0
0.01
0.1
80
90
100
Figure 3-2: Examples of Autoranging Schemes
3.7
The Analyze Function
Normally, all of the functions automatically switch back to the Analyze
function when they have completed their assigned operations. The Escape key
in many cases also switches the analyzer back to the Analyze function.
The Analyze function screen shows the impurity concentration in the first
line, and the range in the second line. In the lower right corner, the abbreviation
Teledyne Analytical Instruments
Part I
3-27
3 Operation
Model 6600
Anlz indicates that the analyzer is in the Analyze mode. If there is an * before
the Anlz, it indicates that the range is linearized.
1.95 ppm
SO2
R1:Ø —10 *Anlz
If the concentration detected is overrange, the first line of the display blinks
continuously.
3.8
Programming
CAUTION: The programming functions of the Set Range and
Curve Algorithm screens are configured at the factory to the users application specification. These functions should only be reprogrammed by trained,
qualified personnel.
To program, you must:
1. Enter the password, if you are using the analyzer’s password
protection capability.
2. Connect a computer or computer terminal capable of sending an
RS-232 signal to the analyzer RS-232 connector. (See chapter 2
Installation for details). Send the rp command to the analyzer.
3. Enter the System menu.
DIG_FILT SELF-TEST
PWD LOGOUT MORE
Use the UP/DOWN switch to blink MORE, then Enter.
AUTOCAL HMGNZR
CAL-HOLD-TIMER
HOLD
MORE
Select MORE and ENTER one more time
ALGORITHM
MODEL OUTPUT:
APPLICATION
4MA
Now you will be able to select the APPLICATION and ALGORITHM
set-up functions.
3-28
Part I
Teledyne Analytical Instruments
Oil in Water Analyzer
Operation 3
3.8.1 The Set Range Screen
The Set Range screen allows reprogramming of the three analysis ranges
and the calibration range (background gas, low end of range, high end of range,
and % or ppm units). Original programming is usually done at the factory according to the customer’s application.
Note: It is important to distinguish between this System programming subfunction and the Range button function, which is an
operator control. The Set Range Screen of the System function allows the user to DEFINE the upper and lower limits of a
range AND the application of the range. The Range function
only allows the user to select or define the limits, or to select
the application, but not to define the application.
Normally the Model 6600 is factory set to default to manual range selection, unless it is ordered as a single-application multiple-range unit (in which case
it defaults to autoranging). In either case, autoranging or manual range selection
can be programmed by the user.
In the autoranging mode, the microprocessor automatically responds to
concentration changes by switching ranges for optimum readout sensitivity. If the
upper limit of the operating range is reached, the instrument automatically shifts
to the next higher range. If the concentration falls to below 85% of full scale of
the next lower range, the instrument switches to the lower range. A corresponding shift in the DC concentration output, and in the range ID outputs, will be
noticed.
The autoranging feature can be overridden so that analog output stays on a
fixed range regardless of the contaminant concentration detected. If the concentration exceeds the upper limit of the range, the DC output will saturate at 1 V dc
(20 mA at the current output).
However, the digital readout and the RS-232 output of the concentration
are unaffected by the fixed range. They continue to read beyond the full-scale
setting until amplifier saturation is reached. Below amplifier saturation, the
overrange readings are accurate UNLESS the application uses linearization over
the selected range.
To program the ranges, you must first perform the four steps indicated at
the beginning of section 3.8 Programming. You will then be in the second
System menu screen.
ALGORITHM
MODEL
APPLICATION
OUTPUT: 4MA
Use the UP/DOWN switch again to move the blinking to APPLICATION
and Enter.
Teledyne Analytical Instruments
Part I
3-29
3 Operation
Model 6600
Sel rng to set appl:
—> Ø1 Ø2 Ø3 <—
Use the UP/DOWN switch to increment/decrement the range number to
01, 02, 03, or CAL, and Enter.
Gas Name
**********
FR:Ø
TO:1Ø %
Use the UP/DOWN switch to increment the respective parameters as
desired, and Enter to move to the next. On the last field %/ppm Enter to accept
the values and return to range selection menu. (See note below.) Repeat for each
range you want to set.
Note: The ranges must be increasing from low to high, for example,
if Range 1 is set to 0–10 % and Range 2 is set to 0–100 %, then
Range 3 cannot be set to 0–50 % since that makes Range 3
lower than Range 2.
Ranges, alarms, and spans are always set in either percent or ppm units, as
selected by the operator, even though all concentration-data outputs change
from ppm to percent when the concentration is above 9999 ppm.
Note: When performing analysis on a fixed range, if the concentration rises above the upper limit as established by the operator
for that particular range, the output saturates at 1 V dc (or 20
mA). However, the digital readout and the RS-232 output
continue to read regardless of the analog output range.
To end the session, send:
st
st
to the analyzer from the computer.
3.8.2 The Curve Algorithm Screen
The Curve Algorithm is a linearization method. It provides from 1 to 9
intermediate points between the ZERO and SPAN values, which can be normalized during calibration, to ensure a straight-line input/output transfer function
through the analyzer.
Each range is linearized individually, as necessary, since each range will
usually have a totally different linearization requirement.
To linearize the ranges, you must first perform the four steps indicated at
the beginning of section 3.8 Programming. You will then be in the second
System menu screen.
3-30
Part I
Teledyne Analytical Instruments
Oil in Water Analyzer
Operation 3
3.8.2.1 Manual Mode Linearization
To linearize manually, you must have previous knowledge of the nonlinear
characteristics of your gases. You enter the value of the differential between the
actual concentration and the apparent concentration (analyzer output). TAI has
tabular data of this type for a large number of gases, which it makes available to
customers on request. See Appendix for ordering information. To enter data:
From the System Functions Screen—
1. Select ALGORITHM , and Enter.
2. Select and Enter SETUP.
3. Enter MANUAL from the Calibration Mode Select screen.
Dpt
Ø
INPUT
Ø.ØØ
OUTPUT
Ø.ØØ
The data entry screen resembles the verify screen, but the gas values can
be modified and the data-point number cannot. Use the UP/DOWN key to
modufy the input field, then Enter to modify the output field, Enter again to move
to the next data field.
After each point is entered, the data-point number increments to the next
point. Moving from the lowest to the highest concentration.
Dpt
0
INPUT
Ø.ØØ
OUTPUT
Ø.ØØ
Repeat the above procedure for each of the data points you are setting (up
to nine points: 0-8). Set the points in unit increments. Do not skip numbers. The
linearizer will automatically adjust for the number of points entered.
When you are done, ESCAPE. The message, Completed. Wait for
calculation, appears briefly, and then the main System screen returns.
To end the session, send:
st
st
to the analyzer from the computer.
3.8.2.2 Auto Mode Linearization
To linearize in the Auto Mode, you must have on hand a separate calibration gas for each of the data points you are going use in your linearization. First,
Teledyne Analytical Instruments
Part I
3-31
3 Operation
Model 6600
the analyzer is zeroed and spanned as usual. Then, each special calibration gas,
for each of the intermediate calibration points, is flowed, in turn, through the
sensor. As each gas flows, the differential value for that intermediate calibration
point is entered from the front panel of the analyzer.
Note: The span gas used to span the analyzer must be >90% of the
range being analyzed.
Before starting linearization, perform a standard calibration. See section
4.4. To enter data:
From the System Functions screen—
1. Select ALGORITHM , and Enter.
2. Select and Enter SETUP.
3. Enter AUTO from the Calibration Mode Select screen.
The Auto Linearize Mode data entry screen appears.
19.5 ppm
SO2
Input(Ø) :20.00
5. Use the UP/DOWN switch to set the proper value of calibration
gas, and Enter. Repeat this step for each cal-point number as it
appears in the Input (x) parentheses.
6. Repeat step 5 for each of the special calibration gases, from the
lowest to the highest concentrations. Escape when done.
To end the session, send:
st
st
to the analyzer from the computer.
3-32
Part I
Teledyne Analytical Instruments
Part I: Control Section
Maintenance 4
Maintenance
Aside from normal cleaning and checking for leaks at the gas connections, routine maintenance is limited to replacing filter elements and fuses,
and recalibration.
WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS
MANUAL.
4.1
Fuse Replacement
The 6600 requires two 5 x 20 mm, 4 A, T type (Slow Blow) fuses.
The fuses are located inside the main housing on the Electrical
Connector Panel, as shown in Figure 4-2. To replace a fuse:
1. Disconnect the Unit from its power source.
2. Place a small screwdriver in the notch in the fuse
holder cap, push in, and rotate 1/4 turn. The cap will
pop out a few millimeters. Pull out the fuse cap and
fuse, as shown in Figure 4-1
Teledyne Analytical Instruments
Part I: 4-1
4 Maintenance
Model 6600 Oil in Water Analyzer
3.0 A MAX
Figure 4-1: Removing Fuse Block Cap and Fuse from Housing
2. Replace fuse by reversing process in step 1.
4.2
System Self Diagnostic Test
1. Press the System button to enter the system mode.
2. Use the < > arrow keys to move to More, and press Enter.
3. Use the < > arrow keys to move to Self-Test, and press Enter.
The following failure codes apply:
Table 5-1: Self Test Failure Codes
Power
0
1
2
3
OK
5 V Failure
15 V Failures
Both Failed
Analog
0
1
2
3
4-2: Part I
OK
DAC A (0–1 V Concentration)
DAC B (0–1 V Range ID)
Both Failed
Teledyne Analytical Instruments
Part I: Control Section
Maintenance 4
Preamp
0
1
2
3
OK
Zero too high
Amplifier output doesn't match test input
Both Failed
>3
Call factory for information
Detector
4.3
0
OK
1
Failed (open filament, short to ground, no
power.)
2
Unbalance (deterioration of filaments, blocked
tube)
Major Internal Components
All internal components are accessed by unbolting and swinging open
the front cover, as described earlier. The major internal component locations
are shown in Figure 4-2, the cell block is illustrated in Figure 3-2, and the
fuse receptacle is shown in Figure 3-3
The 6600 contains the following major internal components:
•
•
•
Customer Interface PCB (Power Supply on bottom surface)
Preamp PCB (Contains Microprocessor)
Front Panel PCB (Contains Displays)
5 digit LED meter
2 line, 20 character, alphanumeric, VFD display
See the drawings in the Drawings section in back of this manual
for details.
For Optical/Detector Alignments, refer to parts II or III of this manual
Teledyne Analytical Instruments
Part I: 4-3
4 Maintenance
Side View
Model 6600 Oil in Water Analyzer
Open Door
Figure 4-2: Control Section Major Internal Components
To swing open the cover panel, remove all screws.
WARNING: HAZARDOUS VOLTAGES EXIST ON CERTAIN
COMPONENTS INTERNALLY WHICH MAY PERSIST
FOR A TIME EVEN AFTER THE POWER IS TURNED
OFF AND DISCONNECTED.
4-4: Part I
Teledyne Analytical Instruments
Part II: Analysis Unit
OPERATING INSTRUCTIONS
Model 6600
Oil in Water Analyzer
Part II: Analysis Section
of the Control/Analysis Unit
6600C - GP, Rack, Panel (Integral or Remote)
6600Z - GP, Bulkhead (Z-Purged in Div II areas)
(Integral or Remote)
6600X - (X-Proof, 1,1,B, C, D) (Integral or Remote)
Teledyne Analytical Instruments
Part II: i
Model 6600 Oil in Water Analyzer
Table of Contents
1 Operational Theory
1.0 Introduction .................................................................... 1-1
1.1 Method of Analysis......................................................... 1-1
1.2 Optical Bench ................................................................ 1-2
1.3 Photometer Amlifier ....................................................... 1-5
1.4 Automatic Zero System .................................................. 1-6
1.5 System Description ........................................................ 1-7
1.6 Photommeter ................................................................. 1-8
1.6.1 Source Module ........................................................ 1-8
1.6.2 Sample Cell ............................................................. 1-9
1.6.3 Detector Module ...................................................... 1-9
1.7 Sample Systems ............................................................ 1-10
2 Installation
2.1 Unpacking the Analyzer ................................................. 2-1
2.2 Installing & Connecting the Analyzer ............................. 2-1
2.2.1 User Connections .................................................. 2-1
2.2.2 Electrical Power Connections ................................ 2-2
2.2.3 Compressed Air Supply ......................................... 2-2
2.2.4 Pipe Connections .................................................. 2-2
2.2.5 Signal and Alarm Output Connections ................... 2-2
2.2.6 Sample Delivery System........................................ 2-2
2.2.7 Draining the System .............................................. 2-3
2.3 Testing the System ......................................................... 2-3
2.4 Calibration ..................................................................... 2-3
2.4.1 Calibration Fluids .................................................. 2-3
2.4.2 Calibration ............................................................. 2-3
3 Maintenance
3.0
3.1
3.2
3.3
3.4
ii: Part II
Routine Maintenance ..................................................... 3-1
Automatic and Routine Operation .................................. 3-1
System Visual Check and Response Procedure ........... 3-1
Routine Maintenance ..................................................... 3-2
Suggested Preventive Maintenance Schedule .............. 3-2
Teledyne Analytical Instruments
Part II: Analysis Unit
3.5 Service Procedures and Adjustments ............................ 3-3
3.5.1 Electronics ............................................................. 3-3
3.5.2 Power Supply Test Points ....................................... 3-3
3.5.3 Setup of the Signal Processing Front-End Amplifier .. 3-3
3.5.4 Oscilloscope Display of the I to E Converter Output .. 3-4
3.5.5 Balancing the Optics for Equal Light Transmission
with Zero Fluid in the Sample Cell ......................... 3-5
3.5.6 Setup of the Logarithmic Amplifier ......................... 3-6
3.5.7 Inverting Amplifier .................................................. 3-6
3.5.8 Integrated Reference and Measuring Signals ........ 3-7
3.5.9 Battery-Powered Oscilloscope Synchronization Point 3-7
3.6 Interface Board Terminal Strip ........................................ 3-7
Appendix
A-1 Specifications ................................................................ A-1
A-2 Recommended 2-Year Spare Parts List ......................... A-3
A-3 Drawing List ................................................................... A-4
Teledyne Analytical Instruments
Part II: iii
Model 6600 Oil in Water Analyzer
iv: Part II
Teledyne Analytical Instruments
Oil in Water Analyzer
Operational Theory 1
Operational Theory
1.0
Introduction
The Teledyne Photometric Analyzer uses the ultraviolet (UV) absorption principle to detect and continuously measure a component of interest in
a sample stream. The analyzer consists of a single sample cell, chopped beam,
folded optics, dual-wavelength UV process photometer and associated microprocessor based control unit and electronics.
1.1
Method of Analysis
The following description shows the course of optical energy in the
analyzer. The optical energy is emitted from a source lamp in the source
module, passed through the sample cell, and received by the sensor, which
converts the optical energy to pulses of electrical energy. These pulses of
electrical energy are processed further in the detector module.
The result is separate pulses that are compared in the control unit to
reveal the measurable difference between optical absorption of the sample at
a selected wavelength (determined by the measuring optical filter) and a zeroabsorption condition (set by the reference optical filter). The magnitude of that
difference represents the concentration of the component of interest in the
sample.
Teledyne Analytical Instruments
Part II:
1-1
1 Operational Theory
1.2
Model 6600
Optical Bench
Depending on the application, the analyzer comes with one of
the following types of lamps: Deuterium (D), Quartz Iodine (L), or Mercury
(Hg). Energy from the lamp, used as a source, is focused through a sample cell
onto a photo detector. In front of the detector is a motor-driven filter disc
containing two optical filters mounted 180 degrees apart that alternately and
continuously rotate into and out of the light beam. Sample flows continuously
through the sample cell and absorbs optical energy at various wavelengths
depending on its composition.
The analyzer monitors two wavelengths: a measuring wavelength selected where the component of interest has a characteristic absorption
peak and a reference wavelength that provides stability by compensating for
extraneous phenomena such as turbidity, cell window deposits, unequal optical
component aging, etc.
Shown with an Integral General Purpose Control and
Analysis Unit with external folded optical bench and
Sample Cell
1-2 Part II
Teledyne Analytical Instruments
Oil in Water Analyzer
Operational Theory 1
Interconnection Diagram
1.3
Photometer Amplifier
The photo detector converts the photo energy striking it to electrical
energy. The magnitude of the photo energy pulses that strike the detector is
determined by absorbance by the sample and the properties of the optical
filters.
The detector output, which is a sequence of pulses that directly reflect
the photo energy transmitted by the measuring and reference filter, is a
measure of the concentration of the component of interest in the sample. The
difference in energy between the measuring and reference pulse is related
exponentially to the concentration of the component of interest.
The photo detector current output is amplified by a current to voltage (I
to E) converting amplifier, followed by a second amplifier. The gain of the
amplifier can be adjusted to obtain any desired output level.
Teledyne Analytical Instruments
Part II:
1-3
1 Operational Theory
Model 6600
To obtain analyzer options that are linearly related to the concentration
of the component of interest, the output of the I to E converting amplifier is fed
to the input of a logarithmic amplifier, which produces a signal that represents
the logarithm of the output signal of the second amplifier. The output of the
logarithmic amplifier is fed to the input of an inverting amplifier, which acts like
a buffer between log amplifier and switch and inverts the input signal for further
processing.
The output of the inverting amplifier is fed to a magnetically activated
SPDT reed switch, synchronized in such a way that all measuring pulses are
collected on one switch contact and all reference pulses on the other.
The pulses pass through diodes that isolate the integrating networks
from each other. The integrators convert the reference and measuring pulse
energy to a DC level representing them. These reference and measuring DC
levels are applied to the subtracting amplifier in the Control Unit. The output
of the subtractor is a DC voltage linearly related to the concentration of the
component of interest.
From the subtractor, the signal progresses to the analog to digital
converter on the motherboard of the Control Unit.
The microcontroller reads the A to D converter and displays the result
on the front panel.
The procedure to set up the optical bench, the signal processing frontend amplifiers, the standardization of outputs, and alarm systems are described
in separate sections of the manual.
1.4
Automatic Zero System
To compensate for zero drift, which may occur during sampling, the
analyzer zero reading is updated by the Auto-Cal function of the controller. An
electronics timing circuit provides a timing cycle that is user programmable.
The Auto-Zero system is turned off (see chapter 3 section 5). You have
the option of setting the analyzer for one six minute zero cycle during hourly
intervals of time from one to 23 hours, and daily from one to 30 days.
The Auto Zero system compares the present zero reading of the zero
fluid with the zero reading of the zero fluid as it was in the last zero calibration.
When there is a difference, the electronic zero circuit sets the zero reading to
1-4 Part II
Teledyne Analytical Instruments
Oil in Water Analyzer
Operational Theory 1
what it was in the last scheduled zero calibration. This zero reading is set
at zero. The Auto Zero circuit is a digital circuit, which employs a DAC (Digital
to Analog Converter) that can go out of range.
When the threshold cannot be found (oscillation persists), this means
that measuring and reference peak signals as viewed on the oscilloscope at the
output of the second amplifier in the detector module are too far out of balance
on zero fluid. When this occurs, you must initiate optical balancing of the
optical filters for equal light transmission on zero fluid. Measuring and
reference peaks must be within one volt with zero fluid in the cell.
Zero drift may occur in the following cases:
1.
The output source changes or chemical or solid deposits form
on the cell windows, but the application is such that interfering chemicals
(sample background changes) are not a problem. The zero fluid in this case
may be the major component of the sample, void of the component of interest.
For oil in very clean water applications, the Zero fluid can be a hydrocarbon free
air or N2.
2.
The sample may contain chemicals that are not of interest, but
absorb UV energy at the measuring wavelength used for analysis of the
component of interest (for example, oil in water applications). These chemicals
produce a signal that adds to the signal of the component of interest and makes
it inaccurate. The Auto Zero system discriminates the two signals and drives the
interfering signal of the background chemicals below zero on an hourly basis.
The zero fluid in this case is the sample of which the component of interest is
filtered out while the background chemicals are preserved. The Auto Zero
system corrects for background changes on an hourly basis, if the analyzer is set
to Auto-Zero in an hourly basis.
1.5
System Description
The photometric analyzer is constructed for hazardouz area (Models
6600, 6600Z-divII or 6600X-div I) use and is mounted on a BACKPLATE, an
open rack, or in a closed cubicle.
Teledyne Analytical Instruments
Part II:
1-5
1 Operational Theory
1.6
Model 6600
Photometer
The photometer modules are mounted on a BACKPLATE inside a
NEMA Enclosure (See D-71055). Facing the mounted photometer, the source
module is at the right top, the sample module is externally located in the folded
optics loop, and the detector module is on the right bottom. A source power
supply module is placed near the HG source module.
1.6.1 Source Module
Any one of three types of source modules may be used in your system.
For oil in water applications the Source module contains a HG source within
its ellipsoid reflector containing a lens and clamp to focus the lamp energy
through the folded optical train.
The QI (Quartz-Iodine) and D2 (Deuterium Arc) sources are mounted
in the source module which also contains the focusing lens.
The source power supply module provides power to the lamps. The
source power supply module houses the power supply, a connector for an
optional temperature controller to heat the sample cell, and an optional span
filter power supply.
The Quartz-Iodine lamp power supply is a switching regulator that
maintains a constant voltage (5 VDC) across the filament of the lamp. The
lamp is incandescent. Its envelope is filled with a halogen to avoid sputtering
of the filament, blackening the lamp envelope.
The D2 lamp power supply is a combination current and voltage
regulator. It maintains a constant anode current in the D2 lamp and controls
the voltage across the lamp’s cathode (filament).
When power is turned on, relay K1 is activated and applies 10 VDC
across the filaments. After ionization of the Deuterium vapor, the lamp starts
to conduct from cathode (filament) to anode. This causes K1 to deactivate and
the filament voltage drops to 7 VDC, which is the operating voltage. The
voltage from anode to cathode which was 365 V before ionization, drops to
about 60 VDC after ignition. This is the operating voltage. A constant current
of 350 mADC is the anode current.
The Deuterium arc lamp is employed with samples whose component
of interest does not absorb at the high intensity peaks of the HG source
emission spectrum. The Deuterium arc produces a broadband of energy (200
to 400 nanometer) in the UV spectrum.
1-6 Part II
Teledyne Analytical Instruments
Oil in Water Analyzer
Operational Theory 1
The HG (Mercury arc) source and its power supply reside in one
enclosure. A quartz lens focuses the energy into a beam for transmission. A
collecting lens is also used at the exit of the folded optical train to focus the source
energy on to the photodetector.
WARNING: UNDER NO CIRCUMSTANCES SHOULD THE
SOURCE MODULE BE OPEN AND THE LAMP ALLOWED TO OPERATE UNLESS PERSONNEL IN
THE IMMEDIATE VICINITY ARE WEARING UV FILTERING EYE GOGGLES.
1.6.2 Sample Cell
The sample cell rests external to the Control and Analysis electronics
placed between the source and detector modules.
1.6.3 Detector Module
The detector module houses the photo detector, chopper assembly, and
the signal processing stages of the electronics circuitry. The synchronized
chopper motor rotates at 1800 rpm. The detector type found in your analyzer can be identified from the Source module sub-assembly (See. D65306).
The filter wheel that carries the optical filters is marked with (M) for
measuring and R for reference filter. If you remove the filter wheel, you must
align a reference mark on the wheel with a reference mark on the shaft. When
the switch activating disc is removed, align with the marks on the switch plate
and motor mount when you put it back.
The phototube detector PC board contains the I to E converter stage,
second amplifier, logarithmic amplifier, inverter, and first stage of integration.
The solid state detector has its I to E converter stage built in on the detector PC
board. A system with a solid state detector has a second converter PC board
containing the second amplifier, logarithmic amplifier, inverter, and first stage
of integration.
The magnetically activated reed switch is mounted on the motor mount.
Oscilloscope test points are available and are mounted on a bracket inside the
housing for explosion-proof models; test points are available on the outside in
the bottom for general-purpose units. An optional zero and/or span filter is
located in this module also.
Teledyne Analytical Instruments
Part II:
1-7
1 Operational Theory
1.7
Model 6600
Sample Systems
Below is a typical sample systems that deliver to the sample fluid 6600
sample cell for Analysis. Depending on the mode of operation either
sample or calibration gas is delivered.
1-8 Part II
Teledyne Analytical Instruments
Oil in Water Analyzer
Part II: Analysis Unit
Installation
Installation of the Model 6600 Photometric Analyzer includes:
1. Unpacking
2. Mounting
3. Fluid connections
4. Electrical connections
5. Testing the system.
2.1
Unpacking the Analyzer
The analyzer is shipped with all the materials you need to install and
prepare the system for operation. Carefully unpack the analyzer and inspect
it for damage. Immediately report any damage to the shipping agent.
2.2
Installing and Connecting the Analyzer
Without Temperature Control, the system must be installed in an area
where the ambient temperature is not permitted to drop below 32°F freezing nor
rise above 122°F (0-50°C).
Regardless of configuration, the system must be installed on a level surface
with sufficient space allocated on either side for personnel and test equipment
access. Subject to the foregoing, the system should be placed as close to the
sample point as possible and bolted to its supporting surface. A waterproof
mastic should be liberally applied to the under surfaces of all four supporting legs
of the cubicle system before placing it in position and bolting it in place.
2.2.1 User Connections
All user connections are around the periphery of the equipment
panel (or cubicle) and appear in the outline diagram in the back of the manual.
Teledyne Analytical Instruments
Part II: 2-1
2 Installation
Model 6600
2.2.2 Electrical Power Connections
Unless specifically ordered, the standard system requires a supply of 115
VAC, single-phase power. Power connections are made inside the control unit.
Refer to the input-output diagram for more specific information. The electrical
power service must include a high-quality ground wire. A high-quality ground
wire is a wire that has zero potential difference when measured to the power line
neutral.
2.2.3 Compressed Air Supply
The system may require a supply of clean, oil and particulate free air to
drive pneumatically activated valves, create suction (pumping) eductor action
(demand more flow), or for use as zero fluids. In general, a 2 liter/minute supply
of compressed air between 80 to 120 psig is usually sufficient. The air supply
must have far greater capacity when purging of the system or when eductors
ejectors are used (special systems).
2.2.4 Pipe Connections
Refer to Appendix Piping Drawings for information about pipe connections. On special systems, consult the text in the manual that describes your
particular sample system in detail.
2.2.5 Signal and Alarm Output Connections
Signal and alarm output connections are made inside the control unit to
terminal blocks mounted on the interface PC board.
Note: For current outputs, the signal circuit resistance, including
accessory devices, must not exceed 1000 ohms. The alarm
contact circuit must not draw more than 3 amperes at 250 VAC
(non-inductive) or 30 VDC. Refer to the following section.
2.2.6 Sample Delivery System
The sample delivery system should be designed to operate reliably and
must be of large enough capacity to avoid flow stops or bubbles in liquid
samples. A pump is required only if there is insufficient pressure to reliably
supply the sample to the system equipment panel. Do not complicate the
delivery system by adding a pump unless it is absolutely necessary. If a pump
is required, select a type that can handle the sample (corrosion), as well as meet
the area classification and Environmental conditions. Choose a pump that can
also supply sufficient flowrates to meet anticipate flow response times based
upon sample delivery take-off distances, line sizes and pressure drops expected
to and from the analysis system.
2-2: Part II
Teledyne Analytical Instruments
Oil in Water Analyzer
Part II: Analysis Unit
2.2.7 Draining the System
In liquid analysis systems, the system return must terminate back to the
process or a safe area as the sample may be poisonous or corrosive. Olso, the
return pressure must be always sufficiently low enough from the inlet pressure
to maintain proper response times within the system.
2.3
Testing the System
Before plugging the instrument into the power source:
• Check the integrity and accuracy of the fluid connections. Make
sure there are no leaks.
• Check the integrity and accuracy of the electrical connections.
Make sure there are no exposed conductors
• Check that sample pressure is between 3 and 40 psig, according
to the requirements of your process.
NOTE: Special designed systems may require checks under vacuum
or high pressure (consult manual addendum). Consult
commissioning start-up section in the manual addendum.
Warning:
Do not operate the “ultrasonic homogenizer” in the
instrument for more than one (1) minute without a
liquid sample properly flowing through the homogenizer.
Power up the system, and test it by performing the following
operations:
1. Repeat the Self-Diagnostic Test, section 3.3.4, part I
2.4
Calibration
2.4.1 Calibration Fluids
Zero and span fluids must be made by the chemistry lab or certified zero
and span fluids bought from a supplier. The zero fluid must be the major
component of the sample, free from the component of interest.
Teledyne Analytical Instruments
Part II: 2-3
2 Installation
Model 6600
The span fluid must be the major component of the sample mixed with
a small amount of the component of interest. The concentration must be 60 to
80% of the range or the widest range of the instrument (if the instrument provides
more than one range).
2.4.2 Calibration
Refer to Section 3.3.8 section I of the manual to determine how to
manipulate the mode setting. Two calibration methods are available.
1. Calibration with zero and span fluids.
2. Calibration with a span filter.
Method One:
1.
Inject zero fluid and set zero as referred in Part I
2.
Inject span fluid and set the concentration of the span fluid with
the span procedure referred in Part I
Method Two:
1.
Determine the span setting using Method One.
2.
Activate the span filter (as referred in section 3.3.8) Part I
3.
Record the display reading (this is the span filter reading and
must be recorded).
4.
You can calibrate the instrument now with the span filter.
Power up the system, and test it as follows:
1. Repeat the Self-Diagnostic Test.
2-4: Part II
Teledyne Analytical Instruments
Oil in Water Analyzer
Maintenance 3
Maintenance
3.0
Routine Maintenance
3.1
Automatic operation and routine operational duties
The system operates continuously without adjustment. Under normal
conditions, after you program the system for automatic operation, only routine
maintenance procedures are necessary. The most common failure condition is
a temporary interruption of the power serving the instrument. If the power
service is interrupted, the source lamp in the analyzer will restart automatically
as long as there is no defect in the lamp circuit or its starter.
You can detect a lamp off condition with the signal failure alarm circuit,
but you must connect the relay contacts from the alarm to your indicating device.
In addition, you will experience an alarm condition when the cell windows are
extremely dirty or the electronics fail in the detector-converter, log amplifier, or
inverter circuits. When the alarm circuit is powered independently from the
analyzer power source, the alarm circuit is fail-safe and will detect power failure.
A message such as "Cell Fail check the detector signal" might be
displayed if lamp off condition occurs
3.2
System Visual Check and Response Procedure
1.
Verify that the signal failure alarm is not in failure condition.
2.
Verify that the zero and span control setting have not been
disturbed (See Part 1).
Teledyne Analytical Instruments
Part II
3-1
3 Maintenance
Model 6600
3.
Verify that the chart recorder contains a normal display.
4.
Verify that the recorder has a sufficient supply of chart paper and
ink.
3.3
Routine Maintenance
Keep the sample lines and components, including the measuring cell
within the analyzer sample module, free of deposits and leaks. You must
determine the interval between cleaning procedures empirically, because the
duration of time that the system runs without attention is related directly to the
sample’s condition. (Some self-cleaning capability has been incorporated into
the inlet flow pattern (turbalent flowingsweep across cell windows) of the
sample cell design.
3.4
Suggested Preventive Maintenance
Schedule
DAILY (these suggestions are perinent to your particular system
design)
1.
Visually inspect the complete system for obvious defects, such
as leaking tubes or connectors.
2.
Verify that the sample pump (if applicable) is running.
3.
Verify that the signal failure alarm is not in failure condition.
4.
Verify that zero and span settings are correct.
MONTHLY
1.
Examine sample cell windows for accumulation of solids.
Remove and clean as necessary.
2.
Calibrate the system. (Check manually the Zero and Span using
prepared Zero/Span fluids obtained from startup or previous calibration practices).
ANNUALLY
1.
3-2
Check the electronics calibration.
Part II
Teledyne Analytical Instruments
Oil in Water Analyzer
Maintenance 3
2.
Check the UV source.
NOTE: Be sure to wear UV filtering eye goggles.
3.
Check the solenoid valves.
3.5
Service Procedures and Adjustments
3.5.1 Electronics
TAI aligns the system’s electronics. However, you may
need to touch up the circuitry, using the following procedure.
Equipment Required:
Oscilloscope (dual trace is preferred, but not required) To observe
oscilloscope test points switch the vertical input selector of the scope to DC.
Switch to AC to observe the demodulator switch signals.
DVM (Digital Voltmeter)
3.5.2 Power Supply Test Points
Measure +15 volt ±1 volt DC and -15 volt ±1 volt DC on the differential
power supply PC board in the control unit. Refer to the power supply schematic
in the back of the manual to identify the power supply test points, or section 3.6
in this chapter.
3.5.3 Setup of the Signal Processing Front-End
Amplifiers
Fill the sample cell with air or a stable fluid, such that the photo
energy that strikes the detector is constant. A stable fluid is distilled or tap
water. This step may be omitted when the system is stable in its present
state.
Teledyne Analytical Instruments
Part II
3-3
3 Maintenance
Model 6600
If you open the detector module, keep stray light out by covering the
opening with a dense black cloth. If you do not take this precaution, the result
is a misinterpretation of the scope patterns. On general-purpose systems, the
scope test points are in the bottom of the detector module and are accessible
without opening the module.
3.5.4 Oscilloscope Display of the I to E Converter Output
The output of the I to E Converter is observed at the output of the second
amplifier. The objective of this operation is to set up the optical system and the
gain of the second amplifier in such a way that the analyzer keeps operating
within its dynamic range.
Connect the oscilloscope to TP3. The oscilloscope displays the measuring and reference pulses in an alternating pattern. The display is created by the
light passing through the reference and measuring filters as they are brought in
and out of the light beam by the rotating filter wheel. These light pulses are
converted to electronic energy which is amplified and brought to TP2. The base
line represents the blocking of the light beam by the opaque part of the filter
wheel.
To identify which of the pulses is the measuring peak, insert the span
filter (when present) or a piece of flat glass or clear plastic in the light beam. The
peak that becomes the shortest (retracts excessively) is the measuring filter pulse.
In case you cannot set the gain properly, because the peaks are too short,
too tall, or too much out of balance, adjust R2 trimpot on the converter PC board
until you obtain the desired peak height as observed on the scope (usually 8 to
9 volt) for the tallest of the two peaks. Never leave the system operating with
peaks exceeding 10 volts or you may saturate the logarithmic amplifier. You
should not permit oscillations or distortions in the peaks.
3-4
Part II
Teledyne Analytical Instruments
Oil in Water Analyzer
Maintenance 3
3.5.5 Balancing the Optics for Equal Light Transmission
with Zero Fluid in the SAMPLE CELL
The objective of this procedure is to obtain measuring and reference
peak heights as displayed on the oscilloscope that are approximately equal, with
the tallest peaks set at 8 to 9 volts. This must be done with zero fluid in the cell.
(Collect a Zero prepared fluid from the sample system ifor all oil in water
analyzers). See Part III, Section 5.5
The procedure is purely mechanical and consists of adjusting the amount
of light passing through either the measuring or reference filter, never both.
Screens (wire mesh) of varying density are used for this operation and are part
of the small took kit accompanying the instrument.
1.
Observe the oscilloscope and judge if optical balancing is
needed. When the difference is less than 1 volt, balancing is not required. The
tallest of the two peaks should be adjusted to 8 or 9 volts with the gain control
R2 on the detector PC board. When this cannot be done because both peaks are
too short or too long, search for screens mounted in the light path, usually located
in a holder on the light pipe which interconnects the detector and sample module,
and remove or add screens, as necessary.
2.
When balancing is needed, identify the peaks as outlined under
Section
3.
For example, if the reference peak is the shorter one, stop the
filter wheel with your hand and see if screens are located behind the reference
filter. The reference filter is identified by the letter “R” engraved on the filter
wheel.
4.
If screens are found, remove them after taking the filter wheel off
the shaft with the special Allen wrench supplied in the tool kit.
5.
After removal of the screens and remounting the filter, mount the
filter wheel back on the shaft. Position it correctly on the shaft by lining up the
two paint marks on shaft and wheel.
6.
Turn on the instrument and observe the balance on the oscilloscope.
a.
If the reference peak is now too tall, remove the filter wheel and
add a screen of lesser density behind the reference filter. Repeat this procedure
until the peaks are within 1 volt of each other.
Teledyne Analytical Instruments
Part II
3-5
3 Maintenance
Model 6600
b. If the measuring peak is equal to or within 1 volt of thereference peak,
the system is optically balanced and ready for calibration.
c. If the peak is still too short, repeat the procedure, but thistime put a
screen behind the measuring filter to shorten its peak.
7. After the peaks are balanced, adjust the gain control until the tallest
of the two peaks is 8 to 9 volts. The peaks should still be within1 volt of each
other.
8. It is always good practice to operate the analyzer with as low a gain
as possible. Therefore, with the gain control just barely off its stop, once again
remove or add screens in the light path to obtain as high a voltage as possible
without exceeding 9 volts for the highest peak. Read-just gain for 8 to 9 volts.
This concludes the balancing procedure and the instrument is ready for
calibration.
3.5.6 Setup of the Logarithmic Amplifier
The amplifier is inverting and continuously taking the logarithm of the
output signal of the second amplifier. You can observe the output by connecting
the scope probe to TP4.
The correct wave shape has a rounded negative going pulse that is the
signal and a flat-topped positive pulse that depicts saturation of the log amplifier.
You should not permit distortions or oscillations in the rounded peaks.
When the positive going pulse is not flat or is distorted, adjust trimpot R3
only enough to obtain a flat positive pulse. If you over adjust, you may lose part
of the second decade of absorption and affect the accuracy of analysis for high
concentrations of the component of interest where the measuring pulse can
become very short. The log amplifier saturates because the amplifier is
incapable of taking the logarithm of the slightly negative baseline.
3.5.7 Inverting Amplifier
The amplifier is inverting and has a gain of 1. It inverts the output signal
of the logarithmic amplifier and acts as a buffer between the logarithmic
amplifier and the reed switch and integrators. To observe the output of the
inverter, connect the scope probe to TP5. The wave must be a duplicateof that
observed on TP4, except that it is inverted.
3-6
Part II
Teledyne Analytical Instruments
Oil in Water Analyzer
Maintenance 3
3.5.8 Integrated Reference and Measuring Signals
You can observe the reference and measuring signal at the first stage of
integration by connecting the scope probe to TP6 (reference signal) and TP7
(measuring signal) at the detector unit. A dual trace scope is advantageous but
not required for this observation.
The test points’ significance is that they reveal proper switch action. The
display shows a sawtooth pattern that is a charge-discharge of the first capacitor
in the integrating network. This ripple is the AC component of the reference and
measuring signal after the pulses are converted to DC. The sawtooth patterns
must be displayed 180° with respect to each other as viewed with a dual trace
scope. They must both be present.
If one is missing, the switch is not switching. If the sawtooth shows a
broken pattern, the switching action is feeble or irregular. Usually, you can fix
the faulty condition of the switch by slightly changing the switch position.
The action of a bar magnet and a rotating chopper disc activate
themagnetic mercury reed switch. An aluminum motor mounting block houses
a bar magnet. This bar magnet is parallel with the mercury chopper switch.
The chopper disc is a green and black disc mounted on the filter wheel
shaft next to the motor. The disc is composed of both magnetic and nonmagnetic materials. As the shaft rotates, the magnetic portion of the disc shorts
the magnetic flux as it passes between the magnet and the switch. The nonmagnetic portion of the disc enables flux lines from the bar magnet to activate
the mercury switch.
3.5.9 Battery-Powered Oscilloscope
Synchronization Point
Because the line frequency cannot synchronize battery-powered oscilloscopes, use TP8 at the detector module to provide external synchronization.
3.6
Interface Board Terminals Strip
At the bottom of the interface PCB on the Control/Analysis Unit, are
three terminal strips where wiring is distributed to other sections of the
Model 6600 System. Such as AC power for the D2 lamp power supply, DC
Power to the preamplifier, High DC voltage for the photodetector, and
Teledyne Analytical Instruments
Part II
3-7
3 Maintenance
Model 6600
signals to control calibration solenoids and filters. To gain access to this
terminals, the silkscreen cover must be removed. These terminals are wired
in the factory.
WARNING: DANGEROUS HIGH VOLTAGES ARE PRESENT AT
THESE TERMINALS. TRAINED PERSONNEL MUST
REMOVE THE SILKSCREEN COVER ONLY. EXERCISE EXTREME CAUTION.
The first strip terminal has three contacts labeled N, G and H. The
labels stand for Neutral, Ground, and Hot. This is the AC power strip
terminal. It feeds AC power to other components of the Model 6600 System, such as the D2 lamp power supply, heater, and temperature controller
PCB.
The second strip terminal has four contacts labeled SHLD, SIG, GND,
MEAS and REF. This strip terminals are dedicated to the signals coming
from the photodetector amplifier. The labels stand for:
SHLD: Shield. Shield form the preamplifier cable connects to this contact.
SIG GND: Signal Ground. Ground reference for both the measure and the
reference signal.
MEAS: Measure Signal voltage.
REF: Reference Signal voltage.
The third terminal strip has eight contacts labeled -230 VDC, +15 VDC, -15
VDC, COM, SPAN FLTR, SPAN SOL, ZERO FLTR, ZERO SOL. This
strip feeds the high voltage needed on the cathode of the photodetector, DC
power for the photodetector preamplifier, and control signals for the solenoids and filters. The labels stand for:
-230 VDC: This is the negative high voltage fed to the photodetector
cathode, about -230 VDC.
3-8
Part II
Teledyne Analytical Instruments
Oil in Water Analyzer
Maintenance 3
+15 VDC: Power Supply voltage fed to the photodetector preamplifier,
+15 VDC.
-15 VDC: Power Supply voltage fed to the photodetector preamplifier,
-15 VDC.
COM: Common reference to the +/- 15 VDC and the -230 VDC power
supplies.
SPAN FLTR: Span filter signal, AC voltage.
SPAN SOL: Span solenoid signal, AC voltage.
ZERO FLTR: Zero filter signal, AC voltage.
ZERO SOL: Zero solenoid signal, AC voltage.
Teledyne Analytical Instruments
Part II
3-9
3 Maintenance
3-10
Part II
Model 6600
Teledyne Analytical Instruments
Part III: Oil in Water
OPERATING INSTRUCTIONS
Model 6600
Oil in Water
Sample Conditioning
System Operation
Part III: Sample System
X-Proof
Part Number D-
6600 - GP, Rack, Panel (Integral or Remote)
6600Z - GP, Bulkhead (Z-Purged in Div II areas)
(Integral or Remote)
6600X - (X-Proof, 1,1,B, C, D) (Integral or Remote)
Teledyne Analytical Instruments
Part III: i
Model 6620 Oil in Water Analyzer
Table of Contents
1.0 Introduction .................................................................... 1-1
2.0
2.1
2.2
2.3
2.4
2.5
2.6
The Method of Analysis ................................................. 1-2
The Optical Bench ......................................................... 1-2
The Photometer Amlifier ................................................ 1-2
The Automatic Zero System ........................................... 1-4
Piping Schematic - B71046-0 ........................................ 1-4
Zero Correction for Clean Background Stream .............. 1-6
Zero Correction for High Background Stream ................ 1-7
3.0 System Description ........................................................ 1-8
3.1 Photommeter ................................................................. 1-9
3.1.1 Source Module ........................................................ 1-9
3.1.2 Sample Cell ............................................................. 1-9
3.1.3 Detector Module ...................................................... 1-9
3.1.4 Control Unit .............................................................. 1-10
3.3 Electrical Connections ................................................... 1-11
3.4 The Sampling System .................................................... 1-11
3.4.1 Sample Water Preconditioning System .................. 1-11
3.4.2 Zero Water Preconditioning System ....................... 1-13
3.4.3 The Automatic Sample Cell Cleaning System ....... 1-14
3.5 The Signal Outputs ........................................................ 1-16
3.6 Recorder Requirements ................................................. 1-16
3.7 The Process Alarm System............................................ 1-17
3.8 The Amplifier PCB ......................................................... 1-17
3.8.1 Auto Zero Circuit .................................................... 1-17
3.8.2 Signal Failure Alarm .............................................. 1-18
4.0 Installation ...................................................................... 1-19
4.1 User Connections .......................................................... 1-19
4.1.1 Electrical Power Connections ................................ 1-20
4.1.2 Compressed Air Supply ......................................... 1-20
4.1.3 Sample Connections ............................................. 1-20
4.1.4 Signal & Alarm Output Connections ...................... 1-20
ii: Part III
Teledyne Analytical Instruments
Part III: Oil in Water
4.1.5 Sample Delivery System........................................ 1-20
4.1.6 Safe Vent (Drainage) .............................................. 1-21
5.0 System Start-up/Calibration ........................................... 1-22
5.1 Installation Check .......................................................... 1-22
5.2 Electronics Check .......................................................... 1-22
5.3 Electrical Check ............................................................. 1-22
5.4 Sample Delivery Check ................................................. 1-23
5.5 Preparation for Calibration ............................................. 1-23
5.5.1 Required Calibration Equipment ........................... 1-24
5.5.2 Acquisition of Representative Oil Sample.............. 1-25
5.5.3 Acquisition of Representative Sample Water ............ 1-25
5.5.4 Oscilloscope Display of the I to E Converter Output .. 1-25
5.5.5 Background Signal Level Determination ............... 1-26
5.5.6 Balancing of the Optics for Equal Light Transmission with
Zero Fluid in the Sample Cell ................................. 1-27
5.6 Calibration Fluid Preparation ......................................... 1-29
5.6.1 Zero Fluid Preparation ............................................ 1-29
5.6.2 Span Fluid Preparation .......................................... 1-29
5.6.3 Calibration ............................................................. 1-30
5.6.4 Calibration by Correlation with Laboratory Analysis1-32
5.6.5 Calibration of Homogenizer ................................... 1-33
5.7 System Set-Up for Automatic Operation......................... 1-35
5.7.1 Set-Up for Automatic Sampling ................................ 1-35
5.7.2 Electronics Set-Up for Automatic Operation.............. 1-35
6.0
6.1
6.2
6.3
Automatic Operation and Routine Operational Duties ... 1-36
System Visual Check and Response Procedure ........... 1-36
Routine Maintenance ..................................................... 1-36
Suggested Preventive Maintenance Schedule .............. 1-37
7.0
7.1
7.2
7.3
7.4
7.5
7.6
7.7
7.8
7.9
7.10
Service Procedures and Adjustments ............................ 1-38
Set-Up of the Signal Processing Front End Amplifier ..... 1-39
Set-Up of the I to E Converter ........................................ 1-39
Set-Up of the Logarithmic Amplifier ............................... 1-40
The Inverting Amplifier ................................................... 1-40
The Integrated Reference and Measuring Signals ......... 1-41
Sample Cell Maintenance .............................................. 1-41
Sample System Maintenance ........................................ 1-42
Zero Filter Replacement ................................................ 1-42
316SS Zero Filter Cleaning Procedure ......................... 1-43
Lamp Replacement ........................................................ 1-44
Teledyne Analytical Instruments
Part III: iii
Model 6620 Oil in Water Analyzer
7.11 Phototube Replacement ................................................ 1-44
8.0 Troubleshooting ............................................................. 1-45
8.1 The Lamp Refuses to Light ............................................ 1-45
8.2 Water Delivery Problems ............................................... 1-45
8.2.1 The Water Refuses to Flow Through the Tubing ........ 1-45
8.2.2 Sample Pump Failure .............................................. 1-46
8.3 Zero Drift Problems ........................................................ 1-46
iv: Part III
Teledyne Analytical Instruments
Oil in Water
Part III
1.0
Introduction
The Teledyne Oil-in-Water Analyzer utilizes the ultraviolet (UV) absorption principle to detect and continuously measure oil concentration in water. The
analyzer consists of two integrated systems: (1) a single external sample cell,
chopped beam, dual-wavelength UV process photometer and associated control analysis unit and electronics, and (2) a sample system that delivers to the
photometer a sample which represents the true oil content of the stream being
analyzed, or a “zero” fluid of oil-free sample delivered to the photometer at a
preset interval once each hour. This oil-free sample is used to reset the zero
reference point on the recorder.
NOTE: Previously, to differentiate between oil and other organic compounds,
oil was formally defined as any material in the sample stream that could be
extracted by carbon tetrachloride, chloroform, hexane, or petroleum ether.
We now know, however, that our oil in water analysis system correlates
exceptionally well to EPA and marine testing methods based upon a more
realistic definition pertinent to how our system works. That is, the definition
of what is truly oil in water from fossil fuel sources is what can be coarse
filtered (non-dissolved oils) and what can be ultra fine filtered (dissolved oils)
from the process water sample. The above definition is a result of a one year
continuous field evaluation by the EPA during the early 80’s which specified
that the Teledyne oil in water system showed good correlation with results
obtained by EPA method 413.1 (superceded in February 1999, by method
1664A using hexane). The field tests were conducted at primary and
secondary effluent sampling points at a refinery. The investigation determined
calibration curves for process oil versus EPA reference oils and validation of
the calibration and sample measurement process against EPA method 413.1
“oil and grease total recoverable.” The report and evaluation was conducted
by the Environmental monitoring and support laboratory, at EPA Cincinnati,
Ohio. The Teledyne Oil-in-Water Analyzer is designed to operate effectively
within the parameters established by this newer accepted definition of oil.
Teledyne Analytical Instruments
Part III:
1-1
Part III
Model 6600
Analytical accuracy of the equipment is better than 2% when it has been
calibrated with an oil identical to that being measured. Reproducibility of
analysis equals or exceeds that of any known laboratory or analytical method.
When calibrated in a range of 0-10 ppm, changes as little as 0.1 ppm are detected
(1% sensitivity).
2.0
The Method of Analysis
The following description follows the course of an optical beam, emitted
from a source lamp in the SOURCE MODULE, passed through the sample to
be analyzed in the SAMPLE CELL, and received (through optical filters),
converted to pulses of electrical energy, and further conditioned, in the
DETECTOR MODULE. The result is separate pulses which are compared in
the control/analysis unit to reveal the measurable difference between optical
absorption of the sample at a selected wavelength (determined by the
MEASURING optical filter) and a zero-absorption condition (set by the
REFERENCE optical filter). The magnitude of that difference represents the
concentration of the component of interest in the sample.
2.1
The Optical Bench
Energy from a Mercury Line lamp, used as a source, is optically focused
through a folded path through a sample cell onto a photo detector. In front of
the detector is a motor-driven filter disc containing two optical filters mounted
180 degrees apart which alternately and continuously rotate into and out of the
light beam. Sample flows continuously through the sample cell and absorbs
optical energy at various wavelengths in accordance with its composition.
The analyzer monitors two of the wavelengths: a measuring wavelength
selected where the components of interest has a characteristic spectral peek
absorbance, and a reference wavelength (where oil does not absorb) utilized to
provide stability by detecting extraneous phenomena such as turbidity, cell
window deposits, unequal optical component aging, etc. The reference wavelength is also sometimes selected at a point where automatic compensation is
attained for interference from other sample components.
2.2
The Photometer Amplifier
The photo detector converts the photo energy impinging on it to electrical
energy. The magnitude of the photo energy pulses which strike the detector is
related to absorbance by the sample and the properties of the optical filters.
1-2 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
The detector output, which is a sequence of pulses which directly reflect
the photo energy transmitted by the measuring and reference filter, is a measure
of the concentration of the component of interest in the sample. The difference
of the energy in the measuring and reference pulse is exponentially related to the
concentration of the component of interest.
The photo detector current output is amplified by a current to voltage (I to
E) converting amplifier, followed by a second amplifier. The gain can be
adjusted to obtain any desired output level.
To obtain electrical signals which are linearly related to the concentration
of the component of interest, the output of the I to E Converting amplifier is fed
to the input of a logarithmic amplifier, which produces a signal that represents
the logarithm of the output signal of the second amplifier. The output of the
logarithmic amplifier is fed to the input of an inverting amplifier, which acts like
a buffer between log amplifier and switch and inverts the input signal for further
processing.
The output of the inverting amplifier is fed to a magnetically activated
SPDT reed switch, synchronized in such a way that all measuring pulses are
collected on one switch contact and all reference pulses on the other.
The pulses pass through diodes which isolate the integrating networks
from each other. The integrators convert the reference and measuring pulse
energy to a DC level representing them. These reference and measuring DC
levels are applied to the subtracting amplifier. The output of the subtractor is a
DC voltage linearly related to the concentration of the component of interest.
From this point on the signal progresses to the A to D converter, where the
signal is digitized for micro controller. The micro controller performs operation
on the signal such as spanning, zeroing, triggering alarms, etc..
The technique of dual wavelength spectroscopy provides compensation
for such phenomena as turbidity, sediment, algae, cell window coatings,
component aging and other extraneous electro-optical attenuation.
Teledyne Analytical Instruments
Part III:
1-3
Part III
Model 6600
The procedure to set up the optical bench, the signal processing front-end
amplifiers, the standardization of outputs, and alarm systems are described in
separate sections for each access.
2.3
The Automatic Zero System
The sample may contain chemicals which are not oil, but absorb UV
energy at the measuring wavelength used for oil analysis, thereby interfering
with the oil analysis (their signals add to the oil signal, which is the only signal
of interest).
The automatic zero system is included to discriminate the two signals. It
involves an electronics circuit, which drives the signal developed by the
interfering, non-oil, background chemicals below zero on an hourly basis.
The electronics zero circuit works in conjunction with a specially designed
sample system.
The sample, which contains oil and background chemicals, is fed to the
sample return port, where it progresses to the various subsections for enhancement in order to present the sample to the measuring cell in such a way so as to
maintain the highest degree of accuracy for oil measurements.
2.4 From B71046-0 (or customer’s) piping schematic
Each of these subsections with piping flow components are identified
with rectangular dotted lines to indicate their importance based upon a
particular oil measurement application.
For Example, there are 4 basic process sampling considerations:
1.
Required use of a homogenizer/dearator and filter assemblies for high
oil range (>20ppm high background applications).
2.
Required use of filter assemblies for low oil (<20ppm oil, high
background).
3.
Required use of back-flush solenoids for low oil range, very low
background applications
4.
Required use of a pump assembly for low pressure (<10 psig) or no
gravity feed applications).
In general, but not without exceptions, the following applications could
involve items 1 through 4 above or combinations thereof.A. 0-20ppm oil
down to 0-10ppm oil in very clean waters such as steam condensates, cooling waters, clear sea waters: (3 and 4, if no sample pressure or gravity feed
available).
Note: Assume sample inlet contains dissolved oil and is homogeneous.
1-4 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
B.
0-20ppm down to 0-10ppm oil ranges in high background
waters such as off-shore platforms, produced waters, sea water, wastewater,
effluents, ponds, bilge/deballasting treatments, on-board ship applications: (2
and 4, if no sample pressure or continuous gravity feed available. (Note:
Assume sample inlet contains dissolved oil and is homogeneous).
C.
0-50ppm to 0-200ppm oil in high background waters such as
off-shore platforms, produced waters, sea water, wastewater, effluents, ponds,
bilge/deballasting treatments, on-board ship applications, tank farms, fuel
depots, rig-washing decks, etc., (1, 2 and 4 if no sample pressures or gravity
feed available). NOTE: for ranges higher than 200ppm oil a dilution system is
required.
Note: Assume sample inlet contains both dissolved and non-dissolved
oil with non-oil organic background compositions and is representative. It
should also be uniform and kept homogeneous up to the homogenization
step.
NOTE: By adjusting valve, V4 in a position for “F1 only”, filtering
versus “F1, F2” and F3 filtering during the auto-zero functioning selects
whether the customer wishes to measure “Total oil and grease recoverable”
or “non-dissolved oil” only. This becomes advantageous when environmental regulation agencies allow tolerable dissolved oil level compositions in the
waters. Many times, cost savings are realized in clean-up operations.
Oil in Water Piping Diagram (B71046)
Teledyne Analytical Instruments
Part III:
1-5
Part III
Model 6600
Should Teledyne receive no representative sample of water or oil for
testing purposes, Teledyne will not be held responsible for the unsatisfactory
functioning of the analyzer due to sample related nonconformities. If the end
user is unable to provide samples, a spectral scan or a detailed description of the
type(s) of oil(s) present is prefferred. The user is responsible for proper
calibration of the unit, at commissioning stage, against the oil(s) found in the
sample fluid to be analyzed. TAI assumes that the sample background is clear
of strong U.V. absorbers at 254nm, i.e., no aromatic hydrocarbons other than oil
and grease listed and NO Fe+3 or H2S are present. Maximum turbidity allowed
is 1 NTU per 10 ppm oil range up to a maximum range of 0-50 ppm oil (20NTU/
200ppm). When no samples are supplied, TAI will ship unit calibrated with ppm
EPA#2 oil in tap water. In some cases, particularly for very high ranges, a
surfactant such as glycol (non-absorbing, non-interfering) is added to the water
sample which increases the miscibility for the oil to go into solution.
The automatic zero cycle is initiated by a signal from the Model 6600
microprocessor based timer circuit. The timing cycle is 1 hour. The timing cycle
can be modified through the system menu, refer to chapter 3 of the control unit
part of this manual. The zero cycle lasts for about 5 minutes and the sample cycle
lasts about 55 minutes.
The sequential switching from sample to zero is operated by the 6600
control unit, where the zero reading of the output is automatically, hourly
upgraded by the auto zero circuit.
2.5
ZERO CORRECTION FOR CLEAN BACKGROUND
STREAMS (REFERENCE B71441 PIPING)
1. The activation of all solenoid valves in the sample system are performed
by the 6600 timer circuit.
2. Backflush solenoids (sv1, sv2) around the sample cell are used to prepare
a reproducible autozero. These solenoids allow oil free air or nitrogen
(N2), (80-120 psig, 5.62-8.44kg/cm2 g) supply is required to purge out
the cell in reverse flow fashion. This action completely cleans the cell
windows and drys them to a stable reproducible background for the
autozero functioning. Because the process is very clean without impurities the autozeroing primariy corrects for lamp, detector and particulate
(dirt) accumulation that may occur on the cell windows.
3. When the span filter with its solenoid is selected/programmed and used
1-6 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
for correcting any gain in the system, its introduction commences at this
time after the zero has been accomplished. The duration to the calibration
of the auto zero is about 10-15 secs longer. This is considered a full autocal updated function where both zero and span are updated each hour.
The air/N2 backflush causes a great disturbance in the detector preamplifiers. The recorder when used and the ppm oil reading on the control unit
digitial display, however, will not notice it due to the hold action of the
sample and hold circuitry (if the analyzer is configured to “hold” and not
track, as mentioned in section 3.39, part I).
2.6
ZERO CORRECTION FOR HIGH BACKGROUND
STREAMS (REFERENCE B71046-0 PIPING) OR CUSTOMER SPECIFIC PIPING)
1. When the stream contains high background impurities, these must be
cancelled out in the autozeroing each hour due to their possible normal
variances in the process stream. In this case we are measuring “total oil
and grease recoverable” and any non-oil organic background hydrocarbons must be corrected for on a continuous basis. In this way, the filter
assemblies are used wherein the filter labeled (F1, 3 micron) eliminates
the non-dissolved oil in the process steam; while filter(s) labeled (F2, 0.2
to 1 micron) and (F3, 0.2 micron teflon element filter) eliminates over
98% of the dissolved oil species. When only the F1 filter is selected by
V4 (in the up position), the autozeroing will cancel out the dissolved oil
left in the process sample; thereby allowing one to measure only the nondissolved oil (This becomes important for some users who are allowed
low levels of dissolved oil in their process stream). For “total oil and
grease recoverable” the V4 3-way valve is selected (in the down positon)
to choose both F1, F2 and F3 filter assemblies thereby autozeroing for all
oil and grease applications. In this way, the non-oil organic background
is cancelled out on a hourly basis.
During the normal measuring function, sv3 is normally opened allowing
the sample to enter the analysis sample cell for oil analysis. When an
autozero timing signal occurs, sv3 is activated and diverts the process
through the filter assembly(s) to correct the background anomolies
usually on a hourly basis.
(NOTE: During calibration practices, a grab sample of the zero process
fluid is obtained here at the Grab zero/Sample Cal Drain Collection.
Valves, V6 and V7 are carefully opened (Caution:Be aware of any
high pressure that may be in the sample) to collect a suitable zero
Teledyne Analytical Instruments
Part III:
1-7
Part III
Model 6600
sample (selected for total oil or non-dissolved oil applications based upon
valve V4 position).
The zero water may be supplied to the measuring cell by the existing process
available pressure, by a user gravity feed system or by a pump (user supplied
at the take-off) or by TAI either remotely at the takeoff source or within the
analyzer sampling system. Normal measuring mode flowrates (commonly
are 100-1000 liter/minute) will render response times of under 10 seconds
typically for 90% FSD (this does not include any lag-time associated by the
fast loop in the system). Each system may be slightly different depending
upon components used (see B71046 piping), but generally the system is
quite fast and many times can be tailored to meet fast response requirements.
The 6600 timer will now activate the auto zero circuit and correct the meter
and recorder to read the zero level previously set to the true absolute zero for
the process fluid.
If the output of the subtractor remains as it was on the preceding zero cycle,
then no correction of the auto zero circuit is required. This means that no
background change has occurred during the past hour.
When the output of the subtractor is other than the previous zero cycle
value, the Auto Zero circuit would compensate for the difference, resulting in
zero volts at its output. This zero voltage, applied to the input of the Sample and
Hold/Buffer circuitry.
The 6600 then (optionally) can insert a manual or automatic employed span
flag that simulates an upscale reproducible calibration analyte of interest.
This function adds another 10-20 seconds to the zero/span check/correction.
After auto zero/span is taken all solenoid valves are de-energized with the
following consequences:
1. The sample is delivered to the sample cell for the next 55 min. The
instrument has returned to the sample cycle and the analyzer monitors
the sample for oil.
3.0
System Description
The oil-in-water analyzer is generally constructed as either an explosion-proof
6600Z or X Purge or general-purpose (Model 6600) unit, open rack or closed
cubicle mounted. An equipment panel is used to support the analyzer components.
1-8 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
All sample-filtering (fluid) components are located on the same side of the
equipment panel as the electrical components.
3.1
Photometer
The photometer control/analysis unit module is mounted on a back panel.
Modules for General Purpose; I, II, B, C & D; and I, I, B, C & D are available and mounted within sheet metal Nema enclosures as well as Nema
stainless enclosures depending upon customer preference for the intended
environmental areas. These enclosures are either Z-purged or X-purged to
meet hazardous area classifications. In some cases, Teledyne has supplied
non-purged, Division II Oil in Water systems self-certified to FM standard
3600 for non-incendive equipment and ISA S12.12-1994 standard for the
same type equipment (Class I, Div II, B, C, D).
3.1.1
Source Module
The source module contains a mercury-line lamp (the source of UV
energy) located within a parabolic reflector which captures most of the emitted
lamp output energy. A quartz lens is used to focus the energy into a beam for
transmission through the optical path and sample cell before reaching the
detector (PMT).
WARNING:
The lamp is installed within a parabolic reflector. It emits strong UV
radiation. When the module is opened with the lamp on, UNDER NO
CIRCUMSTANCES SHOULD THE OPERATOR VIEW THE LAMP
DIRECTLY OR BE ALLOWED TO OPERATE THE UNIT UNLESS
PERSONNEL IN THE IMMEDIATE VICINITY ARE PROTECTED WITH
SUITABLE UV ABSORBING EYEGLASSES.
3.1.2
Sample Cell
All systems, the Aluminum/CPVC sample cell couples the source and
detector modules together along a folded optical train of the photometer. The
sample cell includes double windows (inner sapphire, outer Quartz), at either
end, to prevent condensate from forming in the optical path.
Teledyne Analytical Instruments
Part III:
1-9
Part III
Model 6600
3.1.3
Detector Module
The detector module contains the phototube detector, chopper assembly,
and the first four stages of the electronics circuitry. The synchronized chopper
motor rotates at 1800 rpm.
The filter wheel which carries the optical filters is marked with an “M” for
measuring and an “R” for reference filter. A reference mark on the filter wheel
must be aligned with a reference mark on the shaft in case the filter wheel is
removed from its shaft. Another reference mark is inscribed on the switch plate
and motor mount for the same reason. The detector printed circuit board holds
the I to E Converter stage, second amplifier, logarithmic amplifier and inverter.
The magnetically activated reed switch is mounted on the motor mount.
Oscilloscope test points are available and are mounted on a bracket inside the
housing for explosion- proof models; test points are available on the outside in
the bottom for general-purpose units.
3.1.4
Control/Analysis Unit
The control/analysis unit contains the majority of the electronics employed by the analyzer as well as operator controls. There are four PCB
assemblies. Three of them are located on the door: the Display PCB, The
Main PCB, and the Amplifier PCB. The Interface PCB is located on the
backplate assembly.
The Interface PCB: This large board is where the customer interconnects output signals, Alarm signals, unit receives its AC power, and holds
the DC power supply for the electronics (+5, +/-15 VDC), as well as the
phototube DC power supply to generate -250 VDC. Valve control signals
are interconnected to this board too.
The Amplifier PCB: This board receives the DC voltage signals of the
Measurement and Reference coming from the detector amplifier. The difference of these signals is amplified on this board. Any electronic zeroing
action occurs on this board too. This board mounts on headers that are
available only for that purpose on the Main PCB.
1-10 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
The Display PCB: This board holds the VFD display and the LED
display, and carries the signals to and from the Operator switches on the door
of the 6600.
The Main PCB: This board is mainly digital. The micro controller,
EPROM, RAM, and RS232 driver, as well as digital drivers for alarms, and
valves are located here. There is an analog section where the ADC, DAC,
0-1 VDC output and 4-20 mA modules are located.
3.3
Electrical Connections
There are no electrical junction boxes in the system; power, signal, failure
alarm, and distribution connections are made to terminal blocks located inside
the control/analysis unit housing. Refer to the Interconnection and Wiring
Diagrams at the rear of the manual for details, and chapter of part 1 of this
manual.
3.4
The Sampling System
3.4.1
Sample Water Preconditioning System
The sample water preconditioning system prepares the sample for analysis
on a continuous basis. The sample is homogenized (when range is above
20ppm Oil) so that any undissolved oil fraction will be uniformly dispersed into
solution along with the dissolved oil fraction of the sample prior to presentation
for analysis. Sample pressures ranging up to 150 psi can be accommodated. The
control manifold has a utility water connection selected by a 3-way valve and
inlet port, so that the sample line can be periodically flushed to clear it of
accumulated debris. The untreated sample flows to a high-shear homogenizer
whose purpose is to break up undissolved oil fractions in the sample into droplets
so small that they literally appear to be in solution. This conversion permits the
energy generated at the 254nm measuring wavelength to be absorbed by oil that
would otherwise appear opaque when presented for analysis.
Output of the homogenizer is delivered under a controlled flow for proper
homogenization of the sample water. The homogenized sample is delivered to
the analysis sample cell or through the cell from the auto-zero filter assembly(s).
During the Sample Analysis Cycle, all solenoid valves are de-energized.
SV3 is N.O. and permits the sample to be delivered to the sample cell.
Teledyne Analytical Instruments
Part III:
1-11
Part III
Model 6600
(CONSULT COMMISSIONING/STARTUP PROCEDURE SECTION IN THE ADDENDUM).
PRESSURIZED INLET SAMPLE DELIVERY
In the case of a high background process stream, the sample is delivered to
the sample cell through the homogenizer/deaerator then SV3 (N.O.) directly
to the measuring sample cell for analysis or to the autozero section filter
assembly section before passing through the sample cell. The return must go
to a 5 psig lower pressure point suitable for maintaining a good dynamic
response of the system as well as providing sufficiently high enough flow to
the sample cell in order to maintain some self-cleaning action minimizing
periodic cleaning of the cell assembly. All standard sample handling components selected to contact the process fluid are rated for up to 150 psig (10.5
kg/cm2g). Any pressures higher than this could cause failures and pose an
unsafe condition. It is the customers responsibility to assure the inlet(s) nor
return point(s) do not exceed 150 psig.
In the case of a clean stream, only sv2 and sv1 are used for zero purposes
and are N.O. to allow sample to return to a lower pressure point. This return
point must be at least 10 psig lower than the inlet takeoff (when no pump is
used); otherwise no flow will occur through the system.
NO PRESSURIZED INLET FOR SAMPLE DELIVERY (PUMP
REQUIRED)
When no sample pressure is available either the customer or TAI will have
provided a suitable pump for the delivery of the process fluid.
GRAVITY FEED SYSTEM (Suitable on clean water application only)
For gravity feed systems, the user must provide sample from a suitable head
so as to continuously deliver sample through the highest internal sample
system point of the TAI sample system based upon its commisioned intended
location. The gravity return point must also be suitably below this high point
so as to render sufficient continuous flow through the system.
CALIBRATION SAMPLE DELIVERY
On each side of the sample cell, v5 (before cell) and v6 (after cell) 3-way
valves (followed by v7 an adjustable safety needle valve to a grab safety
port) are placed so as to allow gravity fed entrance of calibration fluids
prepared with the calibration kit (A48715). Sample is blocked off by these
valves and calibration samples prepared can be poured into the zero/span in
calibration reservoir and allowed to flow through the sample cell while
1-12 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
manually calibrating the analyzer system.
V6
V5
V7
Oil in Water Piping Diagram (simplest)
Homogenizing
Input flowrate to the sample homogenizer module inlet is precisely and
accurately controlled by a mechanical flow controller designed for continuous
duty. Because of the style, construction, and position in the sample system, the
controller solves many of the problems associated with sample handling at the
flowrates dictated by the preconditioning technique. For sample concentration
over 20 ppm total oil homogenization is always required-regardless of the
analyzer used to measure the oil above 20 ppm, most waters are so enriched that
oils do not remain homogenized nor miscible enough to measure accurately.
3.4.2
Zero Water Preconditioning System
Since the analyzer operates at a fixed measuring wavelength, and many
soluble organic compounds absorb to some degree at this wavelength, the
effects of organics other than oil must be eliminated from the analysis. Thus, a
solution is prepared that is essentially free of oil, but retains all the other organic
characteristics of the unconditioned sample. When presented to the analyzer, the
absorption caused by the unknown organic compounds in this oil-free water can
be nullified by eliminating any electronic signal that is generated while the
solution is undergoing analysis.
The nondissolved oil is removed by coarse filtering, and the dissolved oil
removed by fine filtering, the water contains only the non-oil organic fraction of
the effluent stream; i.e., it can be used as a reference or “zero oil” water. During
the zero cycle, all solenoid valves are energized and the zero water is pumped
through SV3 to the sample cell. Since turbidity is ratioed out by the electroTeledyne Analytical Instruments
Part III:
1-13
Part III
Model 6600
optics of the photometer design, the analyzer will measure the zero-oil water as
only the non-oil organic fraction. If a differential measurement is made between
the homogenized sample and the zero-oil water, the difference is the oil content
of the stream. If the zero fluid is then made to read zero, the analyzer readout
is a true indication of the oil content of the stream.
NOTE:
Should the fine filter system be by-passed, the analyzer is then
configured to measure non-dissolved oil only.
Because the non-oil organic background of the effluent stream can vary
with time, a correction for its variation is made once each hour. This ensures a
valid oil measurement.
3.4.3
The Automatic Sample Cell Cleaning System
The backflushing of the process fluid can be designed in and to occur for
various reasons depending upon the application. For example,
Ultra clean process applications for leaks into the like:
1
2
3
steam condensates
process cooling water
boiler return condensates
In these applications, backflushing is used not to keep the cell windows
clean but to flush out the process fluid and supply a reproducible zero background fluid (in this case a gas such as oil free air or Nitrogen). When the
process is ultra clean, a clean oil free gas background serves as an excellent
meduim to autozero the instrument on a hourly basis while correcting for the
optics, source and detector anomolies that occur over time. The backflush
supply required (preferred 80-120 psig, 5.6-8.4 kg/cm2g) must be oil free.
Depending upon the oil measuring range of the instrument, a zero offset
factor is programmed into the electronics zeroing function. This is usually set
up at the factory for N2 versus demineralized water. This “zero offset”
eliminates any bias between the ultra clean process liquid zero fluid which is
difficult to obtain on the users site and the surrogate clean air or N2 zero gas
usually readily available. This small bias usually a small percentage of the
range of the instrument is cancelled out by adding in an opposite bias (opposite zero offset, programmed in after field calibration, where the actual zero
offset is determined between the zero pure process water and the surrogate
oil free air or N2). This then, allows the instrument to perform an autozero
function using air or N2 gas instead of preparing a pure zero steam condensate, etc. The bias (called zero offset) is caused by the difference in refractive
1-14 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
index between the ultra pure clean process zero water and air or N2 gas.
Signal level amplitude changes (usually <+/- 25%) will also occur, but their
ratio differences between gas and liquid remain very close also resulting in
close zero readings between gas and liquid.
Entering the zero offset
Determine the valve of the zero offset after field calibrating the unit using the
process fluid. Zero and span calibrate the unit as instructed in part III, sections 5.5 through 5.6.4. Once done, the zero sample in the cell is emptied
from the cell, dried and the reading recorded. Backflushing out the cell can
be done momentarily by commencing a normal autozero function from the
control unit as follows:
On the front panel of the control/analysis unit, operate the switches in the
following manner:
1 hit or scroll the switch labelled escape/enter to enter
display indicates system
flashing
2 hit or scroll the switch labelled down/up to down once
display indicates span
flashing
hit again to down once
display indicates zero
flashing
3 hit enter once
display indicates select zero
manual or auto
4 hit down or up once if display not reading manual and flashing
5
hit enter once
display indicates “zero off”
and “ppm”
flashing
6 hit up or down to enter in offset value (oposite sign, + or -) determined
above between zero water and air or N2 backflush gas.
7 hit enter to start zero function; unit will
display air
purging flashing
If in auto mode of unit proceed to calibrate the unit with the entered zero
offset value put in.
If in manual mode, proceed to calibrate the unit with the coarse and fine zero
adjustments as previously described in Part I, section 3.4.
Teledyne Analytical Instruments
Part III:
1-15
Part III
Model 6600
Particulate laden, turbid, dirtier waters where algae, bacteria,
etc., could collect on the sample cell windows, such as:
1
refinery effluents
2
ponds
3
stagnated effluents
4
oil chemical separators, etc.
In these applications, in addition to the normal autozero liquid functioning,
the backflushing can be used preceding the liquid autozeroing by momentarily actuating when optioned the two solenoid valves sv1 and sv2 across
the sample cell for about 15 seconds before the zero liquid is allowed to
enter the sample cell for background non-organic compound corrections.
The preset open period of the solenoid valves (approximately 15 seconds) is
sufficient to provide a stream of air to the sample cell inlet port at the beginning of each zero cycle. This jet air evacuates the process water and any
collected residue from the sample cell once each hour, vastly reducing the
maintenance that would otherwise be required to keep the cell free of algae
and other debris. The cell inlet/outlet ports are designed with proper sample
inlet/outlet angles across the sapphire windows (non-scratching) thereby
creating high velocity and turbulence both in the normal measuring liquid
stream mode and also during the backflushing cleaning and/or autozeroing
modes described above.
3.5
The Signal Outputs
The standard signal output of the analyzer is 0-1 volt located on the control unit
Interface Board.
The current outputs of the analyzer are isolated 4-20 mA.. The circuit
adjustment procedures the described on section 3, part 1 of this manual.
3.6
Recorder Requirement
The system requires a recorder with input sensitivity which matches the
specified system signal output.
Instruments with floating (ungrounded) current outputs can be connected to any
current input recorder.
1-16 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
The chart speed must be at least 1 inch per hour.
A stripchart recorder is recommended.
3.7
The Process Alarm System
Refers to the control unit part of this manual for interconnection and
programming.
3.8
The amplifier PCB
This board, C67999, contains a differential amplifier. It will take the
difference between the Measurement and the Reference signals to create the
actual output of the amplifier. There are gain settings that are under the
control of the micro controller in the second and the third stage before
delivering the signal to the ADC on the Main PCB.
3.8.1
Auto Zero Circuit
When the oil-in-water sampling system is in long term operation, a zero
signal drift may be caused by various conditions, among which are physical
changes in sample cell conditions (deposits on sample cell windows, etc.),
source and detector changes and electronics drift (with temperature change,
for example).
Periodic compensation for zero drift is accomplished by electronically
nulling the zero offset, with an equal but opposite signal, while zero fluid is
flowing through the system. Thus, a zero point is obtained, and subsequent
sample measurements will produce a signal representing the difference
between the sample measurement signal and the zero reference signal.
The differential amplifier U4 is to be zeroed by signals fed from the
multiple channel DAC on the Main PCB. There are two signals: A Coarse,
and a Fine adjustments. The coarse adjustment is fed thru J2-4 and the Fine
adjustment is fed thru J2-3. Both signals can swing between 0 to 5 volts. But
they have a different effect on the output of the amplifier due the series
resistor value size of each one. The simplified schematic of the amplifier is
shown below. As it can be seen under ideal conditions, setting both the
Teledyne Analytical Instruments
Part III:
1-17
Part III
Model 6600
Coarse and the Fine adjustments to 2.5 VDC should set the output of the
amplifier to zero provided the Measure and the Reference signals have the
same magnitude.
When the control unit enters the zero mode, the micro controller drops
the amplifier to a low gain. The coarse adjustment is first used to zero out the
amplifier on this gain. When the output of the amplifier is reasonably close
to zero, the amplifier goes to the highest gain. Whatever residual offset was
left from the coarse adjustment is now magnified. Now with the coarse
adjustment fixed, the Fine adjustments tries to bring the output of the amplifier back to zero. When the high gain offset is close to zero, the micro controller freezes the Fine and the Coarse adjustments and proceeds to read the
residual offset with the ADC. It goes through all of the ten gains available,
thus the micro controller stores the offset of each gain, so that later they can
be subtracted from the readings. This period of time is called the Software
zero and it takes about 70 seconds to be finished.
F adj
C adj
M sig
U4
R sig
+2.5
3.8.2 Signal Failure Alarm
System contacts are activated when the reference voltage as measured at the
junction of R3 and R4 of the amplifier board, has fallen below 0.50 millivolts
The VFD display will show the following message:
Detector Fail Check the
Detector signal
1-18 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
The reference voltage is checked by the ADC every ten minutes.
This indicates one or more of the following potential failure conditions:
1. The lamp has failed.
2. The cell windows have become so dirty that no light can pass through
the sample.
3. The sample is opaque, due to the presence of solids or undesirable
optical phenomena in the sample.
4. Condensation has occurred on the cell windows.
5. The signal processing electronics, phototube, signal cable, or demodulator switch has failed.
6. The filter wheel is loose on the motor shaft.
7. The reference optical filter is dirty or defective.
8. The optical bench or lamp is out of alignment.
4.0
INSTALLATION
The oil-in-water system must be installed in an area where the ambient
temperature is not permitted to drop below 32oC or rise above 122oF
(50oC),(for salt water applications, the freezing point is lower due to the salt
content). Cubicle installed systems (Models 6600), subject to the preceding
conditions, may be installed outdoors. Rack-supported systems must be
completely sheltered from the elements.
Note: Ambient temperatures below 0oC, requires heated enclosures.
Rack-supported systems must be installed in a well-ventilated area to
prevent the surrounding atmosphere from becoming saturated with the
moisture being generated by the sample preconditioning processes of the
system.
Only in one case when the sample system uses gravity flow should the
system be installed on a level surface otherwise all other systems ar5e to
be with sufficient space allocated on either side (6 feet minimum) for personnel and test equipment access. Subject to the foregoing, the system should
be placed as close to the sample point as possible and bolted to its supporting
Teledyne Analytical Instruments
Part III:
1-19
Part III
Model 6600
surface. A waterproof mastic should be liberally applied to the under surfaces of all four supporting legs of the cubicle system before placing it in
position and bolting it in place.
4.1
User Connections
All user connections are located around the periphery of the equipment
panel (or cubicle) and are shown in the Outline Diagram.
4.1.1
Electrical Power Connections
The system requires a 1 KW supply of 115 VAC, single-phase power.
Power connections are made inside the control unit. Refer to the Interconnection Diagram. The electrical power service must include a high-quality
ground wire. A high-quality ground is defined as having zero potential
difference when measured to the power line neutral.
4.1.2
Compressed Air Supply
The system requires a supply of compressed air at 80-120 psig.
NOTE: The air supply must be oil free. As a precaution, however,
the air pressure line should contain a filter(s) to remove any traces of oil,
supplied and maintained by the customer.
4.1.3
Sample Connection
The sample water input (see Flow Diagram) is connected at the manual
valve input. A flowrate of 1 gallon/minute, minimum, is recommended.
Maximum flow is governed by the impedance of the delivery system.
4.1.4
Signal and Alarm Output Connections
Signal and alarm output connections are made inside the control unit to
terminal blocks mounted on the interface board. Refer to chapter 2 of
control unit part of the manual.
NOTE: Signal circuit resistance, including accessory devices, must not
exceed 1000 ohms. The alarm contact circuit must not draw more than 3
amperes at 250 VAC (None inductive), or 30VDC.
1-20 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
4.1.5 Sample Delivery System
The sample delivery system should be designed to operate at the full
capacity of the interconnecting pipe. Ideally, both the sample and bypass
valves should be adjusted to maximum so that the only impedance to sample
flow is the length and diameter of the delivery pipe. Such a delivery system
will virtually eliminate problems associated with oil and debris collecting in
the sample line. Air ingestion will also be reduced to a minimum, as will
analysis lag time.
TET/AI recommends that the sample line be constructed of 1/2” schedule 80 PVC pipe capable of pressure to 150 psig maximum. Unless absolutely necessary, do not install any valves or restrictions in the line other than
those required to bypass the customer installed external sample pump. All
control over sample flow should be performed at the system inlet and the
pump if provided be allowed to operate at full efficiency against only the
resistance of the line.
A pump is required only if there is insufficient pressure to lift the
sample to the top of the system equipment panel. Do not complicate the
delivery system by adding a pump unless it is absolutely necessary. A low
pressure system will be prone to sample line depositing, but this can be
alleviated by scheduled flushing with the high pressure utility water.
If a pump is required, TET/AI recommends a total submersion centrifugal type. Sub- merging the pump in the sample water automatically eliminates the most common problem - priming the pump. Also, the sample
should be drawn from a point where there is a minimum of turbulence; in
this way, air or turbid suspended solids will not be ingested along with the
water.
The intake side of the system should be equipped with a coarse screen
filter. However, do not filter the water downstream from the sample point,
or use a fine filter anywhere in the sampling system. High filtration will
prevent a representative analysis of the undissolved oil in the sample water.
The inlet of the pump should be placed at a depth where the best
representative concentration of oil can be obtained. Positioning will vary
with the application; however, as a general rule, to avoid skimming, the inlet
should be about 2 feet beneath the surface of the sample water.
4.1.6
Safe Vent (Drainage)
Teledyne Analytical Instruments
Part III:
1-21
Part III
Model 6600
The system safe vent should be equipped to accept a 1-1/2” drain pipe.
Also, the diameter of the drain should be large enough to carry away the full
capacity of the incoming 1/2” sample line. The system safe vent should be
vented to atmosphere at the panel, but the user’s system will have to include
a trap and or scrubber (if H2S present, etc.) if the effluent is to be discharged
into a sewage system. It must be remembered that the vent to be is at floor
level; flooding of the installation area will result in the event of stoppages, or
when drain levels are not below the system vent connector. Also, note this
vent is used to accept the auto-zero backflashing of the sample and as such
high release pressure, liquid and/or gas velocities will be encountered.
Never obstruct this vent by any human personnel means to eliminate any
injuries.
5.0
SYSTEM START-UP/CALIBRATION
The information contained in the following subsections deals primarily
with the steps necessary before the total system can be used for continuously
monitoring oil in water. This information involves installation checks, electrical
checks, fluid dynamics, and calibration.
Due to variations in water and oils as encountered under field conditions, certain
adjustments must be made to optimize system performance for the specific site.
This includes calibration.
5.1
Installation Check
1. Observe that power, signal and alarm connections are properly made
and that the system is properly grounded. Refer to Installation Section
of manual and to Outline and Interconnection Drawings.
2. Inspect the external sample delivery system, including sample pump
(when applicable), sample take up point, and source of water sample.
3. Confirm that the recorder is of the correct type.
5.2
Electronics Check
1. Check that all PC boards are in place.
2. Turn off ultrasonic homogenizer by rotating the homogenizer potentiometer fully counter-clockwise (OFF). Refer to calibration section
5.6.5 for final adjustment.
3. Switch on electronics and confirm that 15 volt, and -230 volt, 10 volt is
present.
4. With an oscilloscope, check test points and reed switch action with air
or oil-free water in the cell.
1-22 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
5.3
Electrical Check
1. Inspect the source temperature controller in the source module. Voltage
across the heater when ON should be 110 volt AC and close to 0 volt
when OFF. For explosion-proof system measure at TS14-11 and 12.
For general-purpose systems measure at TS6-11 and 12.
2. Observe that the automatic lamp starter has turned on the source. Open
the source module and briefly glance inside. A violet glow must be
visible when the lamp is on.
WARNING! UV RADIATION CAN DAMAGE THE EYES. Never
look directly at the lamp for an extended period of time without the aid of special
UV-attenuating goggles.
3. Momentarily turn on sample pump to confirm that no shipping damage
has occurred. The motor must start.
5.4
Sample Delivery Check
1. Adjust the input V1 and V6 3-way with V2 safely block valve off so
that incoming water is diverted to safety vent. Assure no leaks occur.
2. Open V2 safety valve and assure utility water flows out vent. Caution
against any high pressure releases.
3. Request associate to turn on the external user installed sample pump
(when applicable). Flush the lines until the water is consistent in
appearance through the bypass flowmeter F2. If the water is very dirty,
do not permit it to enter the instrument’s sample system: The water
source must be improved.
4. Once sample bypass flow appears very clean as noticed by the flow
through bypass flowmeter, start homogenizer and open F2 sample
flowmeter and with V2 and V3 normally open flush sample through cell
for 10 minutes and notice oil ppm display. Assure analyzer is stable
before proceeding to calibration (See manual addendum for commissioning/starup procedures).
5.5
Preparation for Calibration
Teledyne Analytical Instruments
Part III:
1-23
Part III
Model 6600
Manual Sample Introduction
Manual sample introduction is sometimes desired for the following reasons:
1. Analyzer calibration.
2. Optical balancing on tap water, sea water, or zero water prepared from
the sample.
Prepare the system as follows:
1. Make sure that the instrument is in the analyze mode. The solenoid
valves are programmed to be deactivated so that the sample is connected to the pump or take-off and sample cell.
2. Turn off the homogenizer.
3. Turn off the sample pump.
5.5.1
Required Calibration Equipment
The following laboratory accessories will be required to properly calibrate
the analyzer:
1. Blender: A Waring Model 1120 (or suitable equivalent) laboratory type
blender will be required to prepare the span calibration fluid.
CAUTION
Do not attempt to use the system homogenizer assembly to prepare the
span fluid. Correct control of the precise volumes required for the proper
preparation of the span fluid cannot be achieved with the system homogenizer.
2. Microliter Syringe. Microliter syringes or pipettes will be required to
prepare the span fluid.
3. Graduate. A 500 milliliter graduate will be required to measure the
precise volume of water used in the preparation of the span fluid.
1-24 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
4. Analytical Filter Paper. A supply of Scheicher & Schuell #588, size 24
cm., fast speed, natural finish, 0.008” thickness filter paper will be
required to properly prepare the sample water for use as zero and span
standardization fluids. Glass fiber filters, such as GF/C from Whatman,
may be used and are superior to paper filters. However, a disadvantage
in using this filter is that a Buchner funnel, aspirator flash, and water
aspirator must be used to pull the sample through the filter under
vacuum. The advantages are that its fine particle retention capabilities
are closely matched to the analyzer filter, and filtration proceeds much
faster than gravity feed through paper.
5. Erlenmeyer Flask. Several 500 milliliter Erlenmeyer flasks will be
required to collect, prepare, and handle the sample water during the
preparation of the standardization fluids.
6. Beakers. At least two (2) one liter beakers will be required while
preparing the zero standardization fluid.
7. Sample Bottles. A number of one (1) gallon bottles will be required to
collect the sample for zero fluid preparation or corroborative lab analysis. The bottles should always be thoroughly cleaned before use.
5.5.2
Acquisition of Representative Oil Sample.
A representative oil must be obtained from the user to be used for
calibration. When this oil is not readily available, skim off some of the oil
floating on the surface of the water treatment tanks and remove water and
solids with a centrifuge. Further dry over anhydrous sodium sulfate and
filter.
5.5.3
Acquisition of Representative Sample Water.
The sample water is important since it may contain compounds other
than oil which absorb at the measuring wavelength, causing background
interference which requires compensation.
Teledyne Analytical Instruments
Part III:
1-25
Part III
Model 6600
5.5.4
Oscilloscope Display of the I to E Converter
Output.
The output of the I to E Converter is observed at the output of the
second amplifier. The objective of this operation is to set up the optical
system and the gain of the second amplifier in such a way that the analyzer
keeps operating within its dynamic range, despite variations in turbidity and
changes in background component levels. It is therefore important to measure these components in the sample water.
The “reference” medium against which this data is obtained is either
deionized or distilled water, city tapwater, or pure ocean water in the sample
cell. While the transmittance of the cell filled with different references may
differ, for this purpose these are negligible. Choose the reference that is most
abundantly available on the site, usually tap or ocean water, and use it as the
reference water.
Connect an oscilloscope to TP2. The oscilloscope displays the measuring and reference pulses in an alternating pattern. The display is created by
the light passing through the reference and measuring filters as they are
brought in and out of the light beam by the rotating filter wheel. These light
pulses are converted to electronic energy which is amplified and brought to
TP2. The base line represents the blocking of the light beam by the opaque
part of the filter wheel.
To identify which of the pulses is the measuring peak, insert a piece of
flat glass or clear plastic in the light beam. The peak that becomes the
shortest (retracts excessively) is the measuring filter pulse.
5.5.5
Background Signal Level Determination
1. Draw representative Zero fluid a sample from the grab sample port using
valve V3 of the sample system into a clean(Hydrocarbon free) one
gallon bottle.
2. Collect the sample through filters F1 and F2. This removes nondissolved oils and solid particulates (turbidity). The filtered sample
through both F1 and F2 removes both non-dissolved dissolved oils from
the process fluid. The water is now free of oil and solids, and any
remaining UV absorbance is caused by the non-oil background.
1-26 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
3. This ultra filtered water sample is called zero fluid for the process stream.
a. Gravity feed introduce through the calibration reservoir into the
sample cell this Zero fluid colected above. Record the heights of the
reference and measuring peaks after identification of the peaks.
b. Prepare a Span fluid using the same Aero fluids. See Part III,
Section 5.6.2.
Gravity feed the Span fluid into the calibration reservoir and Sample
Cell. Record meter reading. Record the measuring and reference
peak heights observed on the oscillioscope.
Interpretation of Observations
The upscale reading of the meter in Step b shows the relative signal
level of the oil equivalent signal. The oscilloscope may show a gross imbalance of the peaks as compared to the Zero water peaks.
NOTE: The cell can be charged with zero water (filtered) (or
Span) spaged sample) by pouring the fluid into the calibration reservoir.
5.5.6
Balancing the Optics for Equal Light Trans
mission with Zero Fluid in the Sample Cell
The objective of this procedure is to obtain measuring and reference
peak heights as displayed on the oscilloscope which are approximately
equal, with the tallest of the peaks set at 7 to 8 volts. This must be done with
filtered sample water (zero water) in the cell which contains background
components.
The procedure is purely mechanical and consists of adjusting the
amount of light passing through either the measuring or reference filter,
never both. Screens (wire mesh) of varying density are used for this operation and are part of the small tool kit accompanying the instrument.
1. Observe the oscilloscope and judge if optical balancing is needed. When
the difference is less than 2 volts, balancing is not required. The tallest
of the two peaks should be adjusted to 7 or 8 volts with the gain control
R1 on the detector PC Board. When this cannot be done because both
peaks are too short or too long, search for screens mounted in the light
path, usually located in a holder on the light pipe which interconnects the
detector and sample module, and remove or add screens, as necessary.
Teledyne Analytical Instruments
Part III:
1-27
Part III
Model 6600
2. When balancing is needed, identify the peaks as outlined under 5.5.4.
3. For example, if the reference peak is the shorter one, stop the filter wheel
with the hand and see if screens are located behind the reference filter.
The reference filter is identified by the letter R scribed on the filter wheel.
4. If screens are found, remove them after taking the filter wheel off the
shaft with the special Allen wrench supplied in the tool kit.
5. After removal of the screens and remounting the filter, mount the filter
wheel back on the shaft. Position it correctly on the shaft by lining up
the two paint marks on shaft and wheel.
6. Turn on the instrument and observe the balance on the oscilloscope.
a. If the reference peak is now too tall, remove the filter wheel and add
a screen of lesser density behind the reference filter. Repeat the
procedure until the peaks are within 2 volts of each other.
b. If the measuring peak is equal to or within 2 volts of the reference
peak, the system is optically balanced and ready for calibration.
c. If the peak is still too short, repeat the procedure, but this time put a
screen behind the measuring filter to shorten its peak.
7. After the peaks are balanced, adjust the gain control until the
tallest of the two peaks is 7 to 8 volts. The peaks should still
be within 2 volts of each other.
8. It is always good practice to operate the analyzer with as low
a gain as possible. Therefore, with the gain control just barely
off its stop, once again remove or add screens in the light path
to obtain as high a voltage as possible without exceeding 8
volts for the highest peak. Readjust the gain for 7 to 8 volts.
This concludes the balancing procedure and the instrument is
ready for calibration.
1-28 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
5.6
Calibration with Prepared Sample
Zero and span fluids are prepared from the sample water which will be
ultimately continuously analyzed.
5.6.1
Zero Fluid Preparation
1. Collect one gallon of water from grab sample port by opening V3.
Make sure V4 is position down to collect sample through both F1, F2, and
F3 Zero preparation filters.
2. Filter one liter of collected water to remove total oil.
3. Divide the solution into two 500 ml portions and set aside for
later use.
NOTE: In the case of ultra pure water applications, obtain
an oil free sample (Zero) of the steam condensate fluid, otherwise use demineralized water.
5.6.2
Span Fluid Preparation
1. Pour 500 milliliters of Zero water into the blender’s container. Obtained from collecting sample out the grab sample port above in section 5.6.1
2. Calculate the number of microliters of oil that will be required
to produce a span fluid of full scale concentration when mixed
with the 500 milliliters of Zero water in the blender.
3. Obtain a small representative specimen of oil. Whenever possible,
this specimen
should be recovered from the actual sample water, so that the
calibration will be as representative of the oil composition of the
sample as is possible.
4. Prepare a syringe containing the number of microliters of oil calcu-
Teledyne Analytical Instruments
Part III:
1-29
Part III
Model 6600
lated in Step 2. Carefully wipe off the residue of oil from the tip
and outer body of the needle once the correct volume has been
drawn into the syringe.
5. Run the blender with lid removed at the highest attainable speed
without spillage.
6. Inject the contents of the syringe into the water midway between
the center of the vortex and the wall of the blender container.
Make sure oil injection happens under water, or part of the oil will
be thrown against the wall of the blender container.
7. Put the lid on the blender container, bring the blender up to maximum speed, and homogenize the contents for exactly 2 minutes.
8. EMPTY THE CONTENTS OF THE BLENDER, WHICH IS
NOW STABILIZED FOR ACCURATE PREPARATION OF
SPAN FLUID. REPEAT THE ENTIRE PROCEDURE, THIS
TIME PREPARED FROM THE COORSE FILTERED AND
FINE FILTERED SAMPLE WATER SET ASIDE AT 5.6.1.4.
5.6.3 Calibration
For the most accurate results, calibration of the analyzer must be done
as soon as the span fluid is made up, to avoid separation of oil and water.
The instrument, which was already prepared for manual calibration
fluid introduction (See Section 5.5) and has been optically balanced on zero
fluid prepared from the user’s sample, is ready for calibration.
Recheck the following before fluid introduction:
1. Analyzer is in the Analyze mode.
2. Air blow down valves are turned off. (Manual Mode Required)
1-30 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
Calibration Fluid Introduction:
1. Introduce zero fluid through the calibration reservoir and allow to
flow through all. This requires V2, V3 open and V4 (adjusted) to
give flow of 50-200ccm as noticed by liquid level dropping in
calibration reservoir for Zero and Span in. See 5.5, part I.
As soon as the sample cell is filled with zero fluid and the display
reading stabilizes, zero the analyzer.
2. As soon as the zero fluid is about to be emptied from the reservoir,
add a small amount of span fluid. Make sure that no air
accidentally enters the sample system. This technique assures a
quick exchange of zero fluid for span fluid and conserves span
fluid. As soon as the sample cell is filled with span fluid and the
display stabilizes for the analyzer span the analyzer (see Part I, Sec
5.5). The instrument is now calibrated.
Setup of Internal Span Flag Calibration
Reintroduce zero fluid through and fill sample cell and stabilize display
again.. As soon as the sample cell is filled with zero fluid and the meter
reading stabilizes at zero again, manually introduce and adjust the span
flagreading according to the procedure outlines in section — of the 6600
Control Unit. The exact ppm value of the span filter (Flag) must be manually introduced first and ppm oil representation determined before the autospan mode can be selected to correct any span errors from the original
calibration automatically.The span flag optical filter has been chosen to give
an approximate upscale reading between 50-90% of the range of the instrument. The span flag sensitivity once setup can be automatically programmed
by the 6600 multiprocessor electronics to update the span sensitivity hourly
during the auto-cal (zero plus span functions) of the instrument measurement.
This enhanced auto-cal feature assures high accuracy is maintained between
normal recommended manual field calibration using zero and span fluids
which must occur within a 6 week period minimum.
Teledyne Analytical Instruments
Part III:
1-31
Part III
Model 6600
Span Filter
This instrument employs a synthetic optical filter that can be operated
manually or automatically which simulates the absorption characteristics of the
required span fluid.
After the analyzer has been zeroed and spanned with known compositions
of the appropriate calibration fluids, the span filter is setup for manual or automatic operation when introduced with a zero prepared fluid in the cell (Note: this
fluid may be clean air or N2 for very clean water stream applications, i.e., clean
condensates) and the appropriate equivalent absorption recorded. Use this
setting for all subsequent auto calibrations involving the span filter.
Sample System
Please refer to all pertinent sections of the 6600 Oil in Water manual for
operation of the Oil in water analyzer and sample system.
The sample system and analysis system are described in this manual.
5.6.4
Calibration by Correlation with Laboratory
Analysis
Because of the nature of the analysis, and the unique manner in which it
is accomplished, laboratory confirmation is recommended, particularly where
initial calibration was accomplished with oil not actually recovered from the
sample water.
The laboratory analytical method employed must be capable of detecting both dissolved and undissolved oil fractions in the sample. We recommend EPA method 413.1 (Replaced by method 1664A, gravemetric using
hexane extracting solvent) for sampling and analyzing as mentioned in
section 1.0 of the introduction.
Meaningful determination of the system’s capabilities requires precise
acquisition and processing of grab samples. Adhere to the following procedures:
1. Insure that the system is functioning correctly as described
earlier in this manual.
2. Do not attempt to grab samples when the analyzer is indicating
drastic changes in oil concentration.
3. Draw two 1-gallon samples from the port for grab sample
opening V4 valve. Use clean containers, and fill alternately, not
1-32 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
individually, to insure that oil concentrations are identical in
each. Two samples are used for duplicate analysis which are
then averaged.
4. One minute after samples are drawn, record the meter reading,
time, and date.
5. Do not allow samples to settle for any great length of time to
avoid excessive adhesion of undissolved oil to the container walls.
6. Analyze both samples in the laboratory.
7. Average the results and plot versus the time and date the samples
were taken. Also plot the systems oil reading on the same chart
at the corresponding time.
Plot both the system and laboratory results twice a day, for four or five
days, on the same graph. If the system is operating properly, and the analyzer is correctly calibrated, the system and laboratory graphs should be
displaying very similar averages. If the system is operating properly, but the
analyzer is off calibration, the system and laboratory graphs should be
trending in the same directions, but the average of the system graph will be
consistently higher or lower than the average laboratory results.
If the analytical results are consistently high or low, a corrected span
setting can be derived mathematically. Note the average system and laboratory results. Calculate the percentage error existing between the two results.
Note, the span setting of the control unit (refer to part I, section 3), use the
calculated percentage error to calculate the new span setting.
Calculate the correction to be made to the span setting:
1. Average the chemical lab reports.
2. Average the analyzer results.
3. When the lab results differ from the analyzer results, correct the span
setting:
(Lab Results) X (Span Setting)
Analyzer Results
4. Readjust the automatic zero reference point setting immediately after
the change in span setting.
See the correlation table below as an example of how analyzer results
and lab results are correlated.
Teledyne Analytical Instruments
Part III:
1-33
Part III
Model 6600
5.6.5
Calibration of the ultrasonic
homogenizer
1
Assure that the instrument has been calibrated normally. See sections 5.5 through 5.6.4.
2
Obtain a 1 to 2 liter representative grab process sample (from a steady
state process condition) (ultrasonic homogenizer is “off”-power potentiometer fully counter-clockwise) from the safe vent port by operating
valves V3 and V4. (Caution: be aware of any high process sample
pressure when collecting the sample out of this vent).
3
Blend this process sample (500 cc) for 2 minutes in the waring blender
from the calibration kit similarly to the procedure given in the calibration
section 5.5 through 5.6.4.
4
once blended, pore the sample thorough the normal calibration fluid train
to the safe atmospheric vent collection and record the ppm oil reading.
5
Within a short time period of a few minutes assuming the process oil
concentration has not changed, adjust homogenizer as follows to the
same steady state process sample reading as obtained for the calibration
reading of the blended sample. This must be done at the prescribed
factory sample flow rate setting.
Refer to the ultrasonic homogenizer sample flow rate setting set up at the
factory and recorded in the addendum section of this manual for the
proper calibration and oil response of the instrument.
6
adjust and record the “new” ppm oil reading for this power level potentiometer setting (also recorded) to obtain the same ppm oil reading of the
steady state process sample recorded above in item 4.
The homogenizer is now power level calibrated to give the same ppm oil
response as the calibration system while calibrated for the process sample
flowrate response through the ultrasonic homogenizer.
1-34 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
Warning: Do not operate the “ultrasonic homogenizer” in the
instrument for more than one (1) minute without a
liquid sample properly flowing through the
homogenizer.
Homogenizer Tuning
Follow the procedure in the manual for the homogenizer set up. The final
factor power settings obtained for a 200 ccm flow rate on this system (EPA#2
oil in demineralized water) was estimated at 600 pot dial or 60% power
(approx. 50 watts being used). To arrive at the field calibrated power homogenizer value based upon the unit being calibrated with the customer crude oil in
process water, the homogenizer will have to be recalibrated in the field since the
power setting will be different for the customers oil and water than for what TAI
calibrated it for using EPA#2oil in demineralized water at the 200 ccm flow rate.
A representative collected process sample is pumped through the analyzer
that has been field calibrated using the cal kit blender, etc., (p/n A48715) as
outlined in this manual. This collected sample of process water is again blended
for two minutes the same as the calibration samples and entered into/through the
calibration reservoir through the sample cell. The reading is noted/recorded and
thereafter, the homogenizer power level on the same process fluid (within a short
time period from a steady state sample coming in) is tuned to the same reading
as the blended sample reading recorded value. The homogenizer instrument
power level is again recorded for this 6600 unit. Note the flowrate through the
homogenizer is also recorded. (It should typically be recorded while the flow is
controlled at 200ccc for any future calibrations. (This flow rate will give the most
acceptable response times through the system).
CORRELATION
5.7
System Set-Up For Automatic Operation
5.7.1
Set-up for Automatic Sampling
After calibration with prepared samples, the system must be prepared
for automatic sampling.
1. Insure that sample delivery system was checked per Section 5.4.
2. Close the utility water supply valve used for external sample
line flashing.
Teledyne Analytical Instruments
Part III:
1-35
Part III
Model 6600
3. Close the safety block valve V4.
4. Open the sample inlet valve V1.
5. Open the bypass flowmeter F1 fully.
6. Open sapmle the input header supply valve fully.
7. Start the user’s external sample pump, when applicable.
8. Adjust the sample flowmeter until a flowrate through the bypasses
and sample 5lpm/200ccm respectfully.
9. Switch the pump on, if applicable.
10. Switch the homogenizer on. (verify sample flow is 200ccm)
11. Insure that analyzer power is on and the lamp is on.
12. Make sure (or set) that the analyzer is in the analyze mode.
23. Inspect for proper pump performance. No bubbles should be
visible in the pump tubing.
5.7.2 Electronics Set-up for Automatic Operation
The 6600 control unit timer should have been set on the factory. The
instrument will do its first Automatic calibration three minutes after the
instrument is in the Analyze mode after power up. If this does not happen, or
you wish to confirm the timer setup, or wish to change the set up, check the
AUTO-CAL function in the System menu of the control unit. Refer to
chapter 3 of the control unit part of this manual.
6.0
AUTOMATIC OPERATION AND ROUTINE
OPERATIONAL DUTIES
The system is designed to operate continuously without adjustment. Under
normal conditions, once the system has been programmed for automatic
operation, only routine maintenance procedures are necessary. Perhaps the
most common failure encountered is a temporary outage of the power serving
the system. If the power service is interrupted, the source lamp of the analyzer
will restart automatically, as long as no defect has occurred in the lamp circuit
and its starter. A lamp off condition can be detected by the Signal Failure Alarm
circuit; the relay contacts that are switched by the circuit must be connected to
the customer’s indicating device. In addition, an alarm condition is indicated
when the cell windows are extremely dirty or an electronics failure has occurred
in the Detector-Converter, Log Amplifier or Inverter circuit. When the alarm
circuit is powered independently from the analyzer power source, the alarm
circuit is fail-safe and will detect power failure.
1-36 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
6.1
System Visual Check and Response Procedure
1. Make sure that the signal failure alarm is not in the alarm condition.
2. Check the sample pump operation and function, if applicable
3. Check to see that the homogenizer is running.
4. Check the recorder chart for a normal display, if applicable
5. Make sure that the recorder will not run out of chart or ink.
6.2
Routine Maintenance
The sample lines and components, including the measuring cell within the
analyzer sample module, must be kept free of algae, debris, and undue quantities
of deposited oil to insure accurate analysis.
The interval between cleaning procedures must be determined empirically,
since the duration of time that the system will run without attention is directly
related to the sample’s condition. The frequency of attention is affected by the
following conditions:
1. If the water has been bacteriologically treated, the growth of algae in the
sample passages of the system is a prime consideration. This will dictate
maintenance frequency or corrective action.
2. If undissolved oil is the predominant component of the sample, deposits
in the system become a consideration, particularly if the concentration
exceeds 50 ppm without the homogenizer, otherwise above 200ppm oil.
3. Debris and particulate matter suspended in the sample water will
increase the need for frequent cleaning, since the sample cannot be
filtered.
4. The zero water filter element may become clogged to the point of flow
restriction, at which time it must be replaced or cleaned thoroughlly if
possible with a detergent solvent. Filter element is polypropylane.
Teledyne Analytical Instruments
Part III:
1-37
Part III
6.3
Model 6600
Suggested Preventive Maintenance Schedule
(Application dependent)
DAILY
1. Visually inspect complete system for obvious defects, such as leaking
tubing or connectors, pump failure, and the like.
2. Insure that sample pump is running and that there are no air bubbles
in tubing.
3. Check that homogenizer is running.
4. Check that signal failure alarm is out of alarm condition.
5. Check for correct span setting.
WEEKLY
1. Check condition of zero filter and clean or replace as necessary.
2. Examine sample cell for dirt or oil film accumulation. Remove and
clean as necessary. Clean and reassemble in a dry area to avoid
condensation.
MONTHLY (or 6 weeks maximum)
1. Flush out sample system to remove dirt and oil. (Use utility water)
2. Replace tubing system if obvious deterioration or contamination ob
served.
THREE MONTHS
1. Remove filter element fromcartridge holder. Wash with detergent
and hot water and rinse thoroughly.
2. Flush out piping in analyzer with water or air.
Do not clean F2, micron fine filter. If contaminated replace with new
one.
3. Backflush main sample line.
1-38 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
4. Replace any pump malfanctioning parts as necessary.
5. Replace any tubing which cannot be properly cleaned.
6. Check calibration. Review correlation results. Adjust span setting
as required. Make minor adjustments only. If serious discrepancies
are encountered, use an oscilloscope to check electronics.
7. Re-configure span flag if used (especially if automatic mode used).
ANNUALLY
1. Check electronics calibration.
2. Perform correlation checks.
3. Check UV source.
4. Check solenoid valves.
7.0
SERVICE PROCEDURES AND
ADJUSTMENTS
The system’s electronics are factory aligned. However, when the need
arises to touch up the circuitry, the following procedure is suggested.
1- Oscilloscope (dual trace preferred but not required).
To observe oscilloscope test points (see Detector Module Assembly Dwg.)
switch the vertical input selector of the scope to DC.
2-DVM (Digital Voltmeter)
The PC Board Extender is used whenever trimpot adjustments must be
made. Because all PC Board connectors are keyed to avoid wrong positioning
in the connectors, the key must be removed and later after testing, reinstalled
with long nose pliers. Turn power off during this operation.
7.1
Set up of the Signal Processing Front
End Amplifiers
Teledyne Analytical Instruments
Part III:
1-39
Part III
Model 6600
Fill the sample cell with air or a stable fluid, such that the photo energy
which strikes the detector is constant. A stable fluid is distilled or tap water, clean
ocean water or filtered and sparged sample. This step may be omitted when the
system is stable in its present state.
Open the detector module for scope probe access to test points in explosion
proof systems. Always keep stray light out by covering the opening with a dense
black cloth. If this precaution is not taken, misinterpretation of the scope pattern
results.
The scope test points on general purpose systems are located in the bottom
of the detector module and accessible without opening the module.
7.2
Set Up of the I to E Converter
The I to E Converter converts the small current pulses, produced by the
phototube to a voltage, the output of this amplifier goes to the input of the second
amplifier which output magnitude can be adjusted by means of a gain control
R2. Its location is on PC Board 1 inside the detector module and on E.P.
Analyzers.
The output of the I to E Converter can be observed, by connecting the
ground lead of the scope to TS11-5 and the scope probe to TS2-1 for explosionproof systems and TS5-1 and TS5-2 for general purpose systems.
The oscilloscope displays the measuring and reference pulses in an
alternating pattern.
The display is created by the light passing through the reference and
measuring filters as they are brought in and out of the light beam by the rotating
filter wheel. The baseline represents the blocking of the light beam by the
opaque part of the filter wheel.
To identify which of the pulses is the measuring peak, insert a piece of flat
glass or Plexiglas and the peak that becomes the shortest (retracts excessively)
is the measuring filter pulse.
In case the gain cannot be properly set due to either too short, too tall or too
much out of balance peaks, refer to Section 1.5. Adjust R2 trimpot on PC1 until
the desired peak height is obtained as observed on the scope, usually 7 to 8 volt,
for the tallest of the two peaks. Never leave the system operating with peaks
exceeding 10 volts or the logarithmic amplifier may saturate. No oscillations or
distortions are permitted on the peaks.
1-40 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
The I to E Converter also has an input offset trimpot R3, which function
is to offset the signal baseline slightly, to clean up the log amplifier outputs signal.
Its adjustment will be covered under Section 7.1.5.
7.3
Set Up of the Logarithmic Amplifier
The amplifier is inverting and continuously takes the logarithm of the
output signal of the second amplifier. The output can be observed by connecting
the scope probe to TS2-3 for E.P. systems and TS5-4 for G.P. systems.
The correct wave shape must have a rounded negative going pulse which
is the signal and a flat topped positive pulse which depicts saturation of the log
amplifier.
No distortions or oscillations are permitted on the rounded peaks. When
the positive going pulse is not flat or is distorted, adjust the offset adjustment
trimpot R3 on the I to E Converter. However, adjust no more than required to
just obtain a flat positive pulse. Over adjusting can result in losing part of the
amplifier’s capability to operate in the second decade of its logarithmic operating
range and will affect the accuracy of analysis for high concentrations of the
component of interest where the measuring pulse can become very short.
Saturation of the log amplifier’s output is due to the amplifiers incapability to
take the logarithm of the slightly negative baseline.
7.4
The Inverting Amplifier
The amplifier is inverting and has a gain of 1. Its function is to invert the
output signal of the logarithmic amplifier and to act as a buffer between the
logarithmic amplifier and the reed switch and integrators. To observe the output
of the inverter, connect the scope probe to TS2-2 for E.P. systems and TS5-5 for
G.P. systems.
The wave shape must be a duplicate of that observed on TS2-2, except it
is inverted.
7.5
The Integrated Reference and Measuring Signals
Teledyne Analytical Instruments
Part III:
1-41
Part III
Model 6600
The reference and measuring signal at the first stage of integration can be
observed by placing the scope probe across capacitors C4 and C5 respectively
located at PC Board 1 in the detector unit. A dual trace scope is advantageous
but not required for this observation.
The test points’ significance is that they reveal proper switch action. The
display shows a sawtooth pattern which is a charge-discharge of the first
capacitor in the integrating network. This ripple is the AC component of the
reference and measuring signal after the pulses are converted to DC. The
sawtooth patterns must be displayed 180 with respect to each other as viewed
with a dual trace scope. They must be both present.
If one is missing, this means that the switch is not switching.
If the sawtooth shows a broken pattern, this means the switching action is
feeble or irregular.
The faulty condition of the switch can usually be corrected by moving the
switch up and down or rotating it in its holder.
The magnetic mercury reed switch is activated by the action of a bar magnet
and a rotating chopper disc.
An aluminum motor mounting block houses a bar magnet. This bar magnet
is positioned in parallel with the mercury chopper switch.
The chopper disc is the green and black disc mounted on the filter wheel
shaft next to the motor. The disc is composed of both magnetic and nonmagnetic
materials.
As the shaft rotates, the magnetic portion of the disc shorts the magnetic
flux as it passes between the magnet and the switch. The nonmagnetic portion
of the disc will allow flux lines from the bar magnet to activate the mercury
switch.
7.6
Sample Cell Maintenance
The Sample Cell needs inspection every 6 weeks. For E.P. systems, open
the sample compartment and inspect the cell windows for dirt and oil film. This
inspection can be quickly done by removing the external folded optyical path
assembly from the main Control/Analysis Unit. Disassemble all, clean and
reassemble exactly as before. Re-zero and span calibrate the unit preferably with
zero prepared water and span fluids. Thereafter, recheck span flow as applicable. Reset span flag valu to new calibration values as performed during startup
procedure.
CAUTION:
wear UV goggles if lamp left on.
1-42 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
Shine a flashlight through the cell.
For G.P. Systems, open control/analysis unit and check optical train for
proper beam alignment from source end to detector input.
Remove the cell assembly. Inspect the cell as described in E.P. Systems.
Disassemble the cell in a dry room where needed and clean the cell with hot
water and detergent.
7.7
Sample System Maintenance
The sample system can be quickly cleaned. All components are connected
to the equipment panel by Swaglok type fitting. Remove all components and
clean with hot water and detergent. It is recommended to renew, rather than
clean the piping which makes up the sample flowpath.
7.8
Zero Filter Replacement
The system is equipped with a polypropylene element when the range is
10 ppm or less. A new filter releases small amounts of signal producing
chemicals, which produce an unacceptable zero reading for this narrow range.
The poly propylene filter(s) element must be cleaned and rinsed of contaminats
before use. Wash thoroughly in dtergent, hot eater, then rinse capiously with
very clean dimeneralized or tap water severa times before re-assembly and use.
Several manual auto-zero’s may be required before a stable reproducible zero
is obtained. The only correct gasket material is Viton 4.
Clean througly or replace the filter element:
1. Stop the sample flow.
2. Open the zero system vent valve. This will vent most of the zero
system water.
3. With an adjustable wrench, open the filter housing.
4. Replace the polypropylene filter element. Clean the element when it not
appear very dirty and reinstall or replace with a spare unit. For cleaning
procedure see Section 7.8.
5. After installation, with the proper care that the gaskets are in place and
the filter housing properly tightened, start the sample flow. Flush the new filter
element by filling and draining the filter housing with sample several times, by
means of the safety vent valves, V4 and V3.
Teledyne Analytical Instruments
Part III:
1-43
Part III
Model 6600
NOTE: To avoid excessive foaming, when a new polypropylene filter is installed, it is recommended to flush the filter with tap
water for 1 hour prior to installation. This removes chemicals used
to manufacture the filter and which cause the foaming.
7.9
316 Zero Filter Cleaning Procedure
Because of its construction, the filtering element may be recovered after use
by employing the proper cleaning procedure. This procedure extends the useful
life of a given element immeasurably. The procedure should be faithfully
adhered to, and is as follows:
1. Remove the filter element from the housing, and the gaskets from the
filter element.
2. Immerse the element in a boiling bath of 15% reagent grade caustic soda
solution for30 minutes. DO NOT FLOW THROUGH THE ELEMENT
AS CAUSTIC PARTICLES COULD BE REMOVED BY THE CLEANING AND FORCED INTO THE FILTER MEDIUM - THUS CLOGGING
THE ELEMENT.
CAUTION: Wear protective goggles and gloves, when dealing with
chemicals such as caustics or acids!
3. Neutralize the boiled filter by immersing it in a 160F bath of 10% Nitric
acid for a period of not over 5 minutes.
4. Immerse the filter in a tank of flowing water and rinse until the pH of the
water is neutral. Again, do not force water through the filter, but allow
the water to overflow from the container holding the filter. Periodically
stop the flow of water and check the pH of the water in the container
holding the filter.
5. After the rinse water is verified as being neutral, remove, drain, and air
dry the element. Store in a plastic container, or place back in service.
1-44 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
7.10 Lamp Replacement
After 1 year operation the lamp may be in need of replacement. The
procedure does not require critical alignment of the optical bench; only an
oscilloscope check on the front end amplifiers is recommended after replacement.
PROCEDURE
1. Turn power off to the control unit.
2. Open the source module.
3. Replace the old lamp.
4. Reassemble components in reverse .
5. Check the oscilloscope test points as described under Section 5.5.4.
NOTE:
Observe eye protection warning as described under Section
3.1.1, final text.
7.11 Phototube Replacement
The phototube has a very long life and is only in need for replacement when
a leak has developed in its quartz envelope, the base or around the connector on
top of the tube, which is almost always caused by rough handling or mechanical
shock.
1. Turn power off to the control unit.
2. Open the detector module.
3. Remove the signal cable.
4. Remove the phototube shield.
5. Replace the phototube.
6. Reassemble the components in reverse order.
7. Check the oscilloscope test point as described under Section 5.5.4.
Teledyne Analytical Instruments
Part III:
1-45
Part III
Model 6600
NOTE: The phototube envelope is very thin quartz for optimum UV
trans mission and therefore extremely fragile. Extreme care must be taken, when
the signal cable is disconnected from its top connector to avoid a leak.
8.0
TROUBLESHOOTING SECTION
8.1
The Lamp Refuses to Light
Possible Causes Remedy
1.
The lamp has failed due to long service.1. Replace the lamp as
per Section 7.10.
2.
The power supply has failed. Remove the power supply from
its mount, located in the source module and replace with a new unit.
8.2
8.2.1
Water Delivery Problems
The Water Refuses to Flow Through the Tubing
This can happens during start-up, after installation of new tubing or when
the system is left dry for a long period of time particularly in gravity fed systems.
The cause may be air trapped in the tubing. The small driving force used to make
the water flow is not able to overcome the resistance offered by an air bubble
trapped inside a dry tube.
Remedy: Force the affected tube to become wet, by opening up the
connector on its down stream side and point the tube downwards to generate a
siphon effect and let the water flow freely for awhile.
The tube sections most often suffering of this phenomena are the tube
connection between the output of the homogenizer and the input of the deaerator
and the tube connection between the zero filter output to the solenoid valve and
from the solenoid valve to the input of the fine filter. The deaerator drain tube
leading to the solenoid valve F1 and SV3 must always have a downward slope.
1-46 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
8.2.2
Sample Pump Failure
Refer to Pump manufacturers recommendations or maintenance.
Repair the above conditions by replacement of the defective component.
Pump trouble reveals itself usually by the malfunction of other sample
system components.
1.
Assume the pump is primed with water, if not prime and design
inlet, go as always to have a suitable head for ease of priming the pump.
2.
Turn on the pump and observe that water flows out of the tube.
3.
Turn off the pump. The flow must come immediately to a stop.
A slight drip of water is permissible but not more than approximately 1 droplet
per 5 seconds.
4.
If the drip rate is unacceptable, check for leaks or faulty connections. Take extra care in installing it.
8.3
Zero Drift Problems
1. When the recorder shows a persistent drift in one direction during the
sample cycle, the drift could be caused by an exceptional long upset in the water
treatment facility; this can be discriminated by operating the system temporarily
with air in the sample cell. Make sure that the output of the I to E Converter
displays peak heights not exceeding 10 volts.
2. Check the quality of both F1, F2, and F3, Zero filters; replace or clean
as necessary.
Teledyne Analytical Instruments
Part III:
1-47
Part III
Model 6600
Commissioning and Start-up
Guide for Oil in Water Analyzer Systems:
Please refer to your particular piping, outline, and
wiring drawings of your supplied system in the addendum
portion of this manual.
General requirements for the oil in water systems are’
Notes:- Sample must be representative and single
phase with no high particulates (<40 NTU) nor
slugs of oil present that could plug inlet lines.
Plugging may require flushing with utility water
and/or disassembly of components for maintenance or cleaning.
- Sampling time delays must fit the process control requirements
A
Site requirements:
1
Protection from the elements: We recommend the
customer provide protection on the Enclosure
(especially on an offshore platform) concerning
the following: i.e., direct sunlight, wind sheltering
with suitable mechanical floor and wall supports
if possible.
Protection (shock mounts) from large vibrations
such as pumps and valve operations.
Always continuous and free flowing with proper
high enough sample inlet gravity fed lines for
prevention of cavitation and easy self-priming of
2
3
1-48 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
4
5
6
pumps. (Some pumps depending upon their
application requirements, may be required to
supply high flow, high differential pressures, high
suction (lift) or head pressures and high absolute
operating pressures). Because of this and other
functions, certain pumps may not be allowed to run
dry. (Consult factory for recommendations on your
particular pump used).
Recommend instrument power source be within+/10% of recommended nominal line voltages and
current.
Recommend spacing be made available for
maintenance and access thereof, i.e., cabinet door
openings, etc.
Sample tap requirements:
a)
Heat tracing to prevent freezing of
water(s), or salt laden produced or rig-wash
sample fluids
b)
Proper gravity fed lines as indicated above
c)
Drain openings or returns for the system
that won’t plug
d)
Sampling probes to be placed for
homogenous and representative extraction without
stagnation or dead-volumes and to prevent
entrained air, suspended solids withdrawal
without using fine filters which could remove oil
being measured.
B
Installation:
1
Electrical connections-
Teledyne Analytical Instruments
Part III:
1-49
Part III
Model 6600
Please refer to the Control Unit Section Part I of the
manual.
a
power
1 utility power (requires remote circuit
breaker operation) for pumps if
elected.. In the event the systems are
turned off for maintenance, etc.(in
particular, a bypass pump,
homogenizer).
2 Sample Utility manually 3-way
solenoid water valve, (selects
between sample or utility waterusually operated manually).
3 Instrument power into control unit
(requires external circuit breaker
operation) for all autozero valves in
sample system, controlled by the
Instrument control unit.
Note: The cell purge solenoids may be
used in a particular application,
instead of zero prepared water.
b
Signal and Alarm connections
1 Please refer to the Reference
drawings below for connections and
test points for the signals and alarms.
2 a)
See Analog Outputs as
indicated under Section electrical
connections.
b)
See also section 1 for the
Analog 4-20 mA Output Calibration,
setup, testing and functioning.
1-50 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
4
c
See section for the Alarms Function
a)
See Alarm Relays, Section
2.2, Part I under electrical connections
DCS connections
1 Please refer to Section 1.5 Control
Unit
Interface Panel for input/output functions related to remote
communications.
2 See Alarm Relays, section 2.2
3 See Digital Remote Cal Inputs for Zero
and Span; section 2.2
Reference drawings
a
Outline drawing (see addendum drawings)
-physical locations and sizes of electrical and
sample connections
b
Interconnect drawing (see addendum
drawings)
-indicates where electrical connections go
3
Sample Connections
a The customer connection to the oil in water analysis
system should enter in a downward position enabling
a constant available gravity fed non-freezing sample
with continuous flows capable of up to 5 liters per
minute at pressure provided.
b Assure there are no blockages nor fine filters at the
sample tap nor beyond up to the oil in water analysis
system.
Teledyne Analytical Instruments
Part III:
1-51
Part III
Model 6600
c Check to see that lines are properly heat traced if
required for ambient temperature extremes and that the
sample will not boil from being too hot (for example:
(Maximum temperature allowed is:52 degrees C (limited
by the plastic filter housings), Internal sample system
operation Pressure limit is:150psig G for produced water).
The inlet sample pressure (457 psig) is reduced by a
forward pressure regulator (PR1) which is followed by a
protecting relief valve in the event of failure of PR1).
d Check and familiarize oneself of the analyzer and sample
system components and to assure the entire piping flow
agrees with the Piping Drawing (see addendum
drawings)
1
Sample connections; Refer to Piping diagram
(see addendum drawings)
a
inlet-sample enters the pump.
b
returns
- sample flows through Cell then exits the
system to a common return
drain
Assure the common drain(s) is always free
flowing and always at atmospheric pressure
(unless back to process by suitable
differential or pumping) and will never
freeze or block (backup) under icy
conditions.
Grab Sample port: Refer to your particular
piping scheme for obtaining the sample for
analysis, etc.OBSERVE EXTREME
CAUTION when operating any needle
c
d
1-52 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
valve(s) and wear protective goggles before
opening grab sample valving as the outlet
could exert pressures to 150 psig at this tap.
4
utility water inlet
a This feature allows flushing of the entire sampling train
including the bypass loop. It is operated manually.
Reference drawings:
a
Piping Diagram, (see addendum drawings)
b
Outline drawing (see addendum drawings)
5
Check for shipping integrity of all components
6
Inspect all connections above for proper installation
recommendations
B Start up of System
On the pump control module if applicable, assure the control
switch is in the off position.
1 Check that:
a the sample pump switch is off
b Sample/utility valve is in the utility position.
2
Apply instrument power— after assurance that all
electrical interconnections are correct. Refer to section
2.3 of the manual “Testing the System”.
3 Preliminarily use gravity feed into pump which is set to the
inlet position, and allow all air, minor particulates, etc., to be bled
through the inlet bypass before the bypass pump is turned on. Once you
are assured of proper sample inlet bypass cleanliness (observe water
clarity of outlet to a drain), proceed as follows:
Teledyne Analytical Instruments
Part III:
1-53
Part III
Model 6600
4 Turn on utility power using circuit breaker switch to systems.
5 Turn on bypass pump when applicable, and assure back-pressure
regulator or recirculation loop around pump is OK for its protection.
Allow 10-30 minutes of flushing inlet bypass while observing the water
outlet to drain cleanliness.
7 Once assured sample inlet is proceed as follows:
a
Switch on V1 using the sample/utility valve set to
utility position with the same V1 open Observe the cleanliness of the
utility water to the drain. (You can again also check water quality at the
sample grab point).
If clear, then turn on Sample pump using sample pump power
(which also turns on the homogenizer when applicable (application
dependent) Allow several minutes for the sample to flush through the
entire measuring train of the system to drain or safe vent.
8 Allow warm-up for 1 hour
The UV lamp in particular requires the most time to stabilize from a cold
start. If the unit has been turned of temporarily for only a few minutes, warm-up
can usually be reduced but observation of the output current or voltage should
be checked for stability of at least +/- 1% full scale range. During the initial
power-on the instrument performs an auto zero (autocal if span flag used) if
AUTOCAL functions are on, so stability should be observed after the hold/
tracking output has been released (typically commences after 3 minutes when
power on). See Section 3.3.9 Hold/Track setup.
C
Testing the system
A
Control Unit
1
Perform a self-diagnostic check of the Control Unit as indicated
under Section 3.3.4 of the manual.
2 Refer to Section 2.3, Part I, for Testing of System on Control Unit.
3 Please refer to Section 3.0 for Setup parameters, operation, programming
of the Control Unit.
B Analysis Unit
1-54 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
1 Refer to Section 2.3, Part II,for Testing of System on Analysis Unit.
C Sample Conditioning System Operation
Please read the conditioning system of part III thoroughly.
1 Starting with the Flow path entering the main sample system after the
initial bypass loop, refer to the piping diagram (B74592) for the following sample
train for the measuring path through the analyzer and to drain:
a Measuring flow train: Sample exits 3 way solenoid valve SV1.
b Auto zero path flow train is: Out of sample header and/or homogenizer
if applicable, splits into the 3 micron Filter, thru F1 and 0.2 micron of F2 if total oil
needed, then into F3 (0.2 micron filter) a continuous bypass filter then on into the
sample cell, Sv1.
c The F2 0.2 micron which removes the dissolved oil. NOTE: Dissolved oil
measuring is an option that the customer may so choose depending upon the need
to measure total oil or only the non-dissolved oil fraction.
d Automatic cell cleaning system is built in into the sample cell design and
the N2 blowback when specific to certain applications.
D
Calibration
1
Review Section 5.0 thru 5.4, then proceed at 5.5 of the manual.
2
Check to assure that the blender is available for sample
homogenization. See 5.5.1.
3
Continue thru sections 5.5.2 thru 5.6.3 using known specific oil
encountered in the process.
4
Follow 5.6.4 for correlation to lab test results if alternate acceptable
lab methods available.
E
Set up of internal calibration span flag
1
See section 5.6.3 after cal fluid introduction. The span flag is then
introduced into the light path of the detection system on top of the
N2 background which resembles the clean steam condensate
stream, etc., The set up procedure is explained in the manual under
3.3.8 of the Control Unit section.
F
Set up for Automatic Sampling
Sample System
Teledyne Analytical Instruments
Part III:
1-55
Part III
Model 6600
1
Follow the standard recommended procedures at indicated under
5.7.1 of the manual.
Electronics
1
Follow the standard recommended procedures at indicated under
5.7.2 of the manual.
G
Automatic Operation and Routine Duties:
of the manual.
See Section 6.0
Potentially a power failure could indicate a lamp failure from the
instruments signal failure alarm. The relay contacts provided that are
switched by the circuit must be connected to the customers indicating
device, (DCS system?). In addition, an alarm condition is indicated when
the cell windows are extremely dirty or an electronics failure has occurred
in the detector-converter, log amplifier or inverter circuit. When the alarm
circuit is powered independently from the analyzer power source, the
alarm circuit is fail-safe and will detect power failure.
Please refer to final system check under Section 6.1 in the manual, Part I.
1-56 Part III
Teledyne Analytical Instruments
Oil in Water
Part III
Teledyne Analytical Instruments
Part III:
1-57
Oil in Water Analyzer
Appendix
Appendix
A-1 Specifications
6600 Digital Control Module:
Ranges: Three Programmable Ranges, field selectable
within limits (application dependent) and Auto
Ranging
Display: 2 line by 20 alphanumeric VFD accompanied
by 5 digit LED display
Signal Output: Two 0-1V DC (concentration and range ID)
Two 4-20mADC isolated (concentration and
range ID)
RS232
Alarm: Two fully programmable concentration alarm
set points and corresponding Form C, 3 amp
contacts. One system failure alarm contact to
detect power, calibration, zero/span and
sensor failure.
Mounting: General Purpose NEMA enclosure with
optional Z or X-Purge for Division II or I
areas, or Cenelec Purge
Operating Temperature: 0-50oC
Teledyne Analytical Instruments
A-1
Appendix
Models 6600
Typical Analytical Performance Specifications:
(will vary per application)
Accuracy: ±2% of full scale possible (Oil in Water)
When calibrated on specific oil of interest.
Noise: Less than ±1%
Drift: Less than ±1% per auto Zero cycle (source/
detector dependent)
Sample Cell: (Aluminum/CPVC for Oil in Water) with
Sapphire/Quartz windows standard. (Kynar,
Kalrez optional)
Cell Length: 1/8 to 3.5” (application dependent)
Flow Rate: 100 ccm to 2 lpm (application dependent)
Light Source: Mercury, Hg
Sensitivity: .08 to 1.0 absorbance units.
Reproducibility: +/-1% of scale or better
Filter Wavelength: 210 to 1000 millimicrons. (application dependant - oil in water 254/365mm).
Sample Pressure: 150 psi maximum (oil in water applications)
Sample temperature: 1-120 oC (34-250 oF) (non freezing, non-boiling)
A-2
Teledyne Analytical Instruments
Oil in Water Analyzer
Appendix
A-2 Recommended 2-Year Spare Parts List
Model 6600
QtyP/NDescription
1
1
1
1
1
5
2
1
2
2
4
2
1
2
2
2
1
1
1
C-67435B
C-67999
D-67990
C-13716
L-269
F-57
F-14
P-43
C87
C128
O52
O51
A-16776
F1295
F9
F68
K101
A48715
S1202
Motherboard, Control Unit
Amplifier, Control Unit
6600 Interface PCB
Detector-Converter PCB
HG Source Lamp
Fuse, 5A Slo-Blo
Fuse, 10A Slo-Blo
Phototube
Sample Cell Window (Quartz)
Sample Cell Window (Sapphire)
Viton O-Ring
Viton O-Ring
Accessory Kit
Fuse, 4A slo-blo
Fuse, 1 amp
Fuse, 3 amp
Repair Kit
Cal Kit
Syringe w/tip
Optional: For when customer has a Filtering autozeroing system.
For high background applications --an autozero fitler system is
needed to cancel out non-oil background organics, etc.
1
2
3
Note:
F1484
F1483
F1538
Element, filter of 3 micron (polypropylene)
Element, filter of <1 micron (polypropylene)
#6750 Filter replacement kit assy for F1537
Filter housing.
Orders for replacement parts should include the part number (if
available) and the model and serial number of the instrument for
which the parts are intended.
Teledyne Analytical Instruments
A-3
Appendix
Models 6600
Orders should be sent to:
TELEDYNE Analytical Instruments
16830 Chestnut Street
City of Industry, CA 91749-1580
Phone (626) 934-1500, Fax (626) 961-2538
TWX (910) 584-1887 TDYANYL COID
Web: www.teledyne-ai.com
or your local representative.
A-3 Drawing List (See manual addendum)
DBDCAA-
Outline Diagram, System
Piping Diagram
Wiring Diagram
Interconnection Diagram
Interconnection Diagram
Schematic - Span Filter
Generic Drawing List
B-36470
B-37533
C-36468
C-65371
B-69728
A-4
Schematic - Detector Module - Phototube
Schematic - Detector - Converter PCB - Phototube
Wiring Diagram - Detector Module - Phototube
Wiring Diagram - Backpanel - Hg Source Module
Wiring Diagram - Span Filter
Teledyne Analytical Instruments