Ametek Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Thermox®
WDG-V, WDG-VC, WDG-VCM
Combustion Analyzer
User Manual
Thermox
PN 9000-133-VE Rev. K
150 Freeport Road
Pittsburgh, PA 15238
Offices
USA - Pittsburgh
150 Freeport Road
Pittsburgh, PA 15238
Ph. 412-828-9040
Fax 412-826-0399
FRANCE
AMETEK Precision Instruments France
Rond Point de l’épine des champs
Buroplus Bat D
78990 Elanbourt
Ph. 33 1 30 68 89 20
Fax 33 1 30 68 89 29
USA - Delaware
455 Corporate Blvd.
Newark, DE 19702
Ph. 302-456-4400
Fax 302-456-4444
GERMANY
AMETEK GmbH
Rudolf-Diesel-Strasse 16
D-40670 Meerbusch
Ph. 49 21 59 91 36 0
Fax 49 21 59 91 3680
USA - Texas
4903 West Sam Houston Parkway North
Suite A-400
Houston, TX 77041
Ph. 713-466-4900
Fax 713-849-1924
MIDDLE EAST – Dubai
AMETEK
PO Box 17067
Jebel Ali Free Zone
Dubai, UAE
Ph. 971-4-881 2052
Fax 971-4-881 2053
CANADA
2876 Sunridge Way N.E.
Calgary, AB T1Y 7H9
Ph. 403-235-8400
Fax 403-248-3550
SINGAPORE
AMETEK Singapore Pte. Ltd.
No 43, Changi South Avenue 2, #04-01
Singapore 486164
Ph. 65-64842388
Fax 65-64816588
CHINA
AMETEK Shanghai (SPL)
Room 408, Metro Tower
#30 Tian Yao Qiao Road
Shanghai 200030
Ph. 86 21 6426 8111
Fax 86 21 6426 7054
INDIA
India Private Ltd.
Plot 148, EPIP Zone Phase 2
Featherlite Tech Park, 1st Floor
Whitefield, Bangalore 560066
Tel: 91-80-6782-3200
Fax: 91-80-6780-3232
AMETEK Beijing
CITIC Building, Room 2305
19, Jianguomenwai Dajie
Beijing 100004
Ph. 86 10 8526 2111
Fax 86 10 8526 2141
BRAZIL
Rod. Eng. Ermênio de Oliveira Penteado, Km 57
Tombadouro, Indaiatuba
Sao Paulo, Brazil 13337-300
Tel: 55 19 3825-8775
Fax: 55 19 3935-8773
© 2014 AMETEK
This manual is a guide for the use of the Thermox WDG-V (VC / VCM). Data herein has been verified and validated and is believed adequate for
the intended use of this instrument. If the instrument or procedures are used for purposes over and above the capabilities specified herein, confirmation of their validity and suitability should be obtained; otherwise, AMETEK does not guarantee results and assumes no obligation or liability.
This publication is not a license to operate under, or a recommendation to infringe upon, any process patents.
Contents
Offices...................................................................................................................... ii
Safety Notes........................................................................................................... vii
Important Notice To Users....................................................................................viii
Environmental Information (WEEE)...................................................................... ix
General Safety Summary......................................................................................... x
Product Damage Precautions.................................................................................. xi
Declaration Of Conformity.................................................................................... xii
Warranty And Claims...........................................................................................xiii
CHAPTER 1 Specifications
Sensor: WDG-V ...................................................................................................1-1
Sensor: WDG-VC ................................................................................................1-2
Sensor: WDG-VCM.............................................................................................1-3
Remote Calibration Unit (RCU)...........................................................................1-5
CHAPTER 2 System Overview
Sensor Operations.................................................................................................2-1
Sensor ..................................................................................................................2-2
Basic Elements of the Sensor.........................................................................2-2
The Oxygen Measuring Cell..........................................................................2-3
Hydrocarbons.................................................................................................2-4
Sensor Operations - WDG-VCM..........................................................................2-5
Close-Coupled Extrative Flue Gas Analyzer........................................................2-5
Detectors...............................................................................................................2-6
Combustibles..................................................................................................2-6
Methane..........................................................................................................2-7
Flow Sensor..........................................................................................................2-7
Overview........................................................................................................2-7
Alarm Limits..................................................................................................2-8
Troubleshooting.............................................................................................2-8
Temperature Sensors.............................................................................................2-9
Cell Temperature Thermocouple....................................................................2-9
Combustible/Box Temperature RTD..............................................................2-9
Heaters..................................................................................................................2-9
Cell Furnace Heater.......................................................................................2-9
Box Heater.....................................................................................................2-9
Electronics..........................................................................................................2-10
I/O ..............................................................................................................2-10
Analog Outputs...................................................................................... 2-11
Alarm Relay Contacts............................................................................ 2-11
Contact Input.........................................................................................2-12
RCU Interface........................................................................................2-12
Communications Interface.....................................................................2-12
Contents | iii
Flame Arrestors...................................................................................................2-13
Start-Up Checklist..............................................................................................2-14
Common Operator Errors...................................................................................2-15
Technical Support...............................................................................................2-16
CHAPTER 3 Installation Guide
Analyzer Location................................................................................................3-2
Inspect Shipping Contents....................................................................................3-2
Mechanical Installation.........................................................................................3-3
Sample Inlet Probe Installation......................................................................3-3
Probe Heater Installation................................................................................3-6
Exhaust Tube Installation...............................................................................3-7
Sensor Mounting............................................................................................3-7
Calibration/Aspirator Air.............................................................................3-13
Z-Purge Connections....................................................................................3-14
Remote Calibration Unit Mechanical Installation.......................................3-18
Wiring.................................................................................................................3-24
General Wiring and Conduit Requirements.................................................3-25
Mains Supply Connections..........................................................................3-26
Current Output Connections........................................................................3-29
RCU Connections...............................................................................................3-31
CHAPTER 4 AMEVision Display User Interface or Communications
Definition of the AMEVision Display User Interface & Purpose........................4-1
AMEVision Display User Interface & Keypad....................................................4-2
Setting the Analyzer Address................................................................................4-3
Connecting to the WDG-V...................................................................................4-4
Connecting a Single Sensor to the AMEVision.............................................4-4
Connecting Multiple Sensors to the AMEVision...........................................4-4
Customer I/O Interconnecting........................................................................4-5
Maximum Distance Between Sensor & AMEVision............................... 4-6
Troubleshooting....................................................................................................4-6
No Communications................................................................................4-6
Intermittent Communications..................................................................4-6
WDG-V PC Configurator Software......................................................................4-7
CHAPTER 5 Flow Setting
CHAPTER 6 LED Status Indicator
iv | Thermox WDG-V / VC / VCM
CHAPTER 7 Troubleshooting
General Troubleshooting......................................................................................7-2
Leak Check....................................................................................................7-2
Plugged Plumbing Check...............................................................................7-3
Aspirator Air Not Pulling Sample from Process............................................7-3
Diagnostics Checks...............................................................................................7-4
Wiring Checks................................................................................................7-5
Interconnecting Wiring Problem....................................................................7-5
Thermocouple Checks....................................................................................7-5
Shorted/Failed Thermocouple........................................................................7-5
Reversed Thermocouple Wires......................................................................7-5
Thermocouple Compensation .......................................................................7-5
Circuit Check.................................................................................................7-5
Calibration/Aspirator Setup Checks...............................................................7-7
AC Power Checks..........................................................................................7-8
Furnace Checks..............................................................................................7-8
Process Pressure Checks................................................................................7-8
Cell Checks....................................................................................................7-8
Alarm Messages (Red Bar on AMEVision Display User Interface)....................7-9
Cell T/C Failure..............................................................................................7-9
Cold Junction Compensator Failure...............................................................7-9
Cell Temperature Control...............................................................................7-9
Cell Over Tem................................................................................................7-9
Cell Under Temp............................................................................................7-9
Cell Temp Rise Failure...................................................................................7-9
Over Temp Relay Tripped..............................................................................7-9
RTD Failure....................................................................................................7-9
Box Temp High..............................................................................................7-9
Box Temp Low...............................................................................................7-9
Box Temp Rise Failure...................................................................................7-9
Low Sample Flow........................................................................................7-10
Cell Failure (Open)......................................................................................7-10
Combustible Detector Open.........................................................................7-10
Methane Detector Open...............................................................................7-10
Cell Mv Mismatch.......................................................................................7-10
Analog Output 1 Failure..............................................................................7-10
Flow Sensor Failure.....................................................................................7-10
Cell T/C Measurement Mismatch................................................................7-10
Contents | v
Warning Messages (Yellow Bar on AMEVision Display User Interface).......... 7-11
Cell Life Nearing its End............................................................................. 7-11
Comb Detector Life Nearing its End........................................................... 7-11
Methane Detector Life Nearing its End....................................................... 7-11
High Cell Mv............................................................................................... 7-11
Analog Output 1 Out of Range.................................................................... 7-11
Analog Output 2 Out of Range.................................................................... 7-11
Analog Output 3 Out of Range.................................................................... 7-11
Last O2 Span Calibration Failed................................................................... 7-11
Last O2 Zero Calibration Failed................................................................... 7-11
Last Comb Calibration Failed...................................................................... 7-11
Last CH4 Calibration Failed........................................................................ 7-11
Last Flow Calibration Failed........................................................................ 7-11
Oxygen Calibration Required...................................................................... 7-11
Combustible Calibration Required............................................................... 7-11
CH4 Calibration Required............................................................................ 7-11
Chapter 8
Service and Parts
Warnings & Cautions............................................................................................8-2
Wiring Diagram....................................................................................................8-3
Cell Replacement/Cleaning..................................................................................8-5
Thermocouple Replacement.................................................................................8-7
Detector Replacement...........................................................................................8-9
Combustibles Detector Replacement.............................................................8-9
Methane Detector Replacement.....................................................................8-9
Flow Sensor Replacement.................................................................................. 8-11
Furnace Replacement.........................................................................................8-13
Heater Replacement............................................................................................8-14
Box Heater Replacement.............................................................................8-14
Combustible Heater Replacement................................................................8-16
Electronics Replacement....................................................................................8-18
Recommended Maintenance Schedule...............................................................8-20
Calibration....................................................................................................8-20
Replacement Parts List.......................................................................................8-21
Chapter 9
Drawings and Custom Instructions
Chapter 10
Customer Modbus Map
Appendix A French Warnings / Français Avertissements
vi | Thermox WDG-V / VC / VCM
Safety Notes
WARNINGS, CAUTIONS, and NOTES contained in this manual emphasize critical
instructions as follows:
An operating procedure which, if not strictly observed, may result in
personal injury or environmental contamination.
An operating procedure which, if not strictly observed, may result in
damage to the equipment.
Important information that should not be overlooked.
NOTE
Burn hazard. Hot surface. Do not touch, allow to cool before servicing.
Electrical Safety
Up to 5 kV may be present in the analyzer housings. Always shut down power source(s)
before performing maintenance or troubleshooting. Only a qualified electrician should
make electrical connections and ground checks.
Any use of the equipment in a manner not specified by the manufacturer may impair
the safety protection originally provided by the equipment.
Grounding
Instrument grounding is mandatory. Performance specifications and safety protection
are void if instrument is operated from an improperly grounded power source.
Verify ground continuity of all equipment before applying power.
| vii
Important Notice to Users
The following applies to the WDG-V Division 2 Models:
Power, input, and output (I/O) wiring must be in accordance with Class I, Division 2
wiring methods [Article 501-4(b) of the National Electric Code, NFPA 70] and in accordance with the authority having jurisdiction.
WARNING - EXPLOSION HAZARD - SUBSTITUTION OF COMPONENTS MAY IMPAIR SUITABILITY FOR CLASS I, DIVISION 2.
WARNING - EXPLOSION HAZARD - WHEN IN HAZARDOUS LOCATIONS, TURN OFF POWER BEFORE REPLACING OR WIRING
MODULES.
WARNING - EXPLOSION HAZARD - DO NOT DISCONNECT EQUIPMENT UNLESS POWER HAS BEEN SWITCHED OFF OR THE AREA
IS KNOWN TO BE NONHAZARDOUS.
THIS EQUIPMENT IS SUITABLE FOR USE IN CLASS I, DIVISION 2,
GROUPS A, B, C, AND D OR NONHAZARDOUS LOCATIONS ONLY.
The maximum ambient temperature for the analyzer is 60°C (140°F).
The WDG-V sensor and AMEVision Display User Interface is a complex piece of
equipment that should only be serviced by a qualified service technician with expertise
in instrument technology and electrical systems. AMETEK recommends that all equipment requiring service be sent back to the factory. You should only attempt to repair or
service this equipment after receiving training from an AMETEK/P&AI Division training representative. If you decide to service this equipment be aware that high voltages,
high temperatures, and other potentially hazardous conditions may arise.
viii | Thermox WDG-V / VC / VCM
Environmental Information (WEEE)
This AMETEK product contains materials that can be reclaimed and recycled. In
some cases the product may contain materials known to be hazardous to the environment or human health. In order to prevent the release of harmful substances into the
environment and to conserve our natural resources, AMETEK recommends that you
arrange to recycle this product when it reached its “end of life”.
Waste Electrical and Electronic Equipment (WEEE) should never be disposed of in a
municipal waste system (residential trash). The Wheelie Bin marking on this product is a reminder to dispose of the product properly after it has completed its useful
life and been removed from service. Metals, plastics, and other components are
recyclable and you can do your part by doing one of the following steps:
• When the equipment is ready to be disposed of, take it to
your local or regional waste collection administration for
recycling.
• In some cases, your “end of life” product may be traded in
for credit towards the pur-chase of new AMETEK instruments. Contact your dealer to see if this program is available in your area.
• If you need further assistance in recycling your AMETEK
product, contact our office listed in the front of the instruction manual.
| ix
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to this
product or any products connected to it.
Use Proper Wiring
To avoid fire hazards, use only the wiring specified in the Installation Chapter of this
user’s manual.
Avoid Electrical Overload
To avoid electrical shock or fire hazard, do not apply a voltage to a terminal that is
outside the range specified for that terminal.
Ground the Product
Follow the grounding instructions provided in the Installation Chapter of this user’s
manual. Before making connections to the input or output terminals of this product,
ensure that the product is properly grounded.
Do Not Operate without Covers
To avoid electric shock or fire hazard, do not operate this product with covers or panels
removed.
Use Proper Fuse
To avoid fire hazard, use only the fuse type and rating specified for this product.
Do Not Operate in Explosive Atmosphere
To avoid injury or fire hazard, do not operate this product in an explosive atmosphere
unless you have purchased options that are specifically designed for these environments.
x | Thermox WDG-V / VC / VCM
Product Damage Precautions
Use Proper Power Source
Do not operate this product from a power source that applies more than the voltage
specified.
Do Not Operate with Suspected Failures
If you suspect there is damage to this product, have it inspected by qualified service
personnel.
Use Caution When Lifting
The analyzer weighs approximately 20 kg (44 lb). Use caution when lifting it from its
crate.
Use Proper Attire
Equippement is hot, user should wear protective groves while handling the equipment.
Contents | xi
Declaration of Conformity
Manufacturer’s Name:
AMETEK/Thermox®
Manufacturer’s Address:
Process & Analytical Instruments Division
150 Freeport Road
Pittsburgh, PA 15238
declares that the products:
Product Names:
Thermox WDG series of Flue Gas Analyzers using AMEVision Display User Interface:
WDG-V WDG-VC
WDG-VCM
Conform to the following standards:
EMC compliance:
EMC Directive 2004/108/EC
Immunity:
EN 50082-2:1995 Generic - Heavy Industrial Equipment
EN 50082-1:1997 Generic - Residential, Commercial, Light Industrial Equipment
EN 61000-6-2:1999 Generic - Heavy Industrial Equipment
EN 61326:1997 Measuring Equipment (Heavy Industrial)
Emissions:
EN 55011:1998 Class A, ISM Equipment
Safety Compliance:
UL 3101-1 UL Listed Inspection and Measuring Electrical Equipment
CAN/CSA 22.2 No. 1010.1-92 cUL Listed Inspection and Measuring Electrical Equipment
Low Voltage Directive 73/23/EEC:
EN61010-1:1993,
Amendment A2: 1995
Electrical Equipment for Measurement, Controls,
and Laboratory Use
Ingress Protection:
IP65
Manufacturer’s Address in Europe:
AMETEK Precision Instruments Europe GmbH
Rudolf-Diesel-Strasse 16
D-40670 Meerbusch, Germany
June 2013
xii | Thermox WDG-V / VC / VCM
WARRANTY AND CLAIMS
We warrant that any equipment of our own manufacture or manufactured for us pursuant to our specifications which shall not be, at the time of shipment thereof by or for us, free from defects in material
or workmanship under normal use and service will be repaired or replaced (at our option) by us free
of charge, provided that written notice of such defect is received by us within twelve (12) months
from date of shipment of portable analyzers or within eighteen (18) months from date of shipment or
twelve (12) months from date of installation of permanent equipment, whichever period is shorter. All
equipment requiring repair or replacement under the warranty shall be returned to us at our factory,
or at such other location as we may designate, transportation prepaid. We shall examine such returned
equipment, and if it is found to be defective as a result of defective materials or workmanship, it shall
be repaired or replaced as aforesaid. Our obligation does not include the cost of furnishing any labor
in connection with the installation of such repaired or replaced equipment or parts thereof, nor does
it include the responsibility or cost of transportation. In addition, instead of repairing or replacing the
equipment returned to us as aforesaid, we may, at our option, take back the defective equipment, and
refund in full settlement the purchase price thereof paid by Buyer.
The warranty shall not apply to any equipment (or part thereof) which has been tampered with or
altered after leaving our control or which has been replaced by anyone except us, or which has been
subject to misuse, neglect, abuse or improper use. Misuse or abuse of the equipment, or any part
thereof, shall be construed to include, but shall not be limited to, damage by negligence, accident, fire
or force of the elements. Improper use or misapplications shall be construed to include improper or
inadequate protection against shock, vibration, high or low temperature, overpressure, excess voltage
and the like, or operating the equipment with or in a corrosive, explosive or combustible medium,
unless the equipment is specifically designed for such service, or exposure to any other service or
environment of greater severity than that for which the equipment was designed.
The warranty does not apply to used or secondhand equipment nor extend to anyone other than the
original purchaser from us.
THIS WARRANTY IS GIVEN AND ACCEPTED IN LIEU OF ALL OTHER WARRANTIES,
WHETHER EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION AND WARRANTIES OF FITNESS OR OF MERCHANTABILITY OTHER THAN AS EXPRESSLY SET FORTH
HEREIN, AND OF ALL OTHER OBLIGATIONS OR LIABILITIES ON OUR PART. IN NO
EVENT SHALL WE BE LIABLE UNDER THIS WARRANTY OR ANY OTHER PROVISION OF
THIS AGREEMENT FOR ANY ANTICIPATED OR LOST PROFITS, INCIDENTAL DAMAGES,
CONSEQUENTIAL DAMAGES, TIME CHANGES OR ANY OTHER LOSSES INCURRED
BY THE ORIGINAL PURCHASER OR ANY THIRD PARTY IN CONNECTION WITH THE
PURCHASE, INSTALLATION, REPAIR OR OPERATION OF EQUIPMENT, OR ANY PART
THEREOF COVERED BY THIS WARRANTY OR OTHERWISE. WE MAKE NO WARRANTY,
EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION ANY WARRANTIES OF FITNESS OR OF MERCHANTABILITY, AS TO ANY OTHER MANUFACTURER’S EQUIPMENT,
WHETHER SOLD SEPARATELY OR IN CONJUNCTION WITH EQUIPMENT OF OUR MANUFACTURE. WE DO NOT AUTHORIZE ANY REPRESENTATIVE OR OTHER PERSON TO
ASSUME FOR US ANY LIABILITY IN CONNECTION WITH EQUIPMENT, OR ANY PART
THEREOF, COVERED BY THIS WARRANTY.
Contents | xiii
This page intentionally left blank.
xiv | Thermox WDG-V / VC / VCM
SPECIFICATIONS
Sensor : WDG-V
Output Range
From 0-1% to 0-100%
Accuracy
± 0.75% of measured value or ± 0.05% oxygen, whichever is greater
Response
90% of a process step change < 6 secs.
Drift
< 0.1% of cell output per month (< 0.005% O2 per month with 2% O2
applied)
Aspirator Air
Requirements
≈ 3 scfh (1.4 L/min) at 15 to 100 psig (1.05 to 7.04 kg/cm2)
Max. Flue Gas Temp / Probe Type / Lengths
1300°F (704°C)
1875°F (1024°C)
3000°F (1649°C)
316 SS
310 SS
Hexoloy®
Max. Sample
Dewpoint
392°F (200°C) standard.
Sample Pressure
± 6 inch water gauge
36” - 108” (91 cm - 271 cm)
36” - 108” (91 cm - 271 cm)
24” - 72” (60 cm-182 cm)
Environment
Ambient Temp.:
-13°F to 149°F (-25°C to 65°C)
-4°F to 140°F (-20°C to 60°C) for Purged Units
Relative Humidity: 10% to 90%, non-condensing
Max. Altitude:
5000 meters
Enclosure
Hinged IP65, Optional Div II
Power Requirements
115 VAC, ± 10%, 47-63 Hz, 600VA
max.
230 VAC, ± 10%, 47-63 Hz, 600 VA
max.
Calibration Gas Requirements
Use calibration gases @ 10 PSIG, 3.0 SCFH (1.4 L/min.)
O2 Span Gas: Air or from 1.0 to 100% O2, balance N2
O2 Zero Gas: 2% or from 0.1 to 10% O2, balance N2
System Compliance
EMC Compliance:
Safety Compliance:
2004/108/EC
73/23/EEC
NOTES
1.
All static performance characteristics are with operating variables constant.
2.
System accuracy referenced to 0.1 to 10% calibrated range.
Specifications |
3.
Response is to calibration gas.
1-1
Sensor : WDG-VC
Output Range
Oxygen
Combustibles
From 0-1% to 0-100%
From 0-1% to 0-100%
From 0-1,000 ppm to 0-10,000 ppm or from 0-1% to 0-5%
Accuracy
Oxygen
± 0.75% of measured value or ± 0.05% oxygen, whichever is greater
Combustibles
± 2% of full scale output range
Response O2
90% of a process step change < 6 secs.
Drift
< 0.1% of cell output per month (< 0.005% O2 per month with 2% O2
applied)
Aspirator Air
Requirements
≈ 3 scfh (1.4 L/min) at 15 to 100 psig (1.05 to 7.04 kg/cm2)
Max. Flue Gas Temp / Probe Type / Lengths
1300°F (704°C)
1875°F (1024°C)
3000°F (1649°C)
316 SS
310 SS
Hexoloy®
Max. Sample
Dewpoint
392°F (200°C)
Sample Pressure
± 6 inch water gauge
36” - 108” (91 cm - 271 cm)
36” - 108” (91 cm - 271 cm)
24” - 72” (60 cm-182 cm)
Environment
Ambient Temp.:
-13°F to 149°F (-25°C to 65°C)
-5°F to 140°F (-20°C to 60°C) with Div. 2 Option
Relative Humidity: 10% to 90%, non-condensing
Max. Altitude:
5000 meters
Enclosure
Power Requirements 115 VAC, ± 10%, 47-63 Hz, 600VA max.
230 VAC, ± 10%, 47-63 Hz, 600 VA max.
Calibration Gas Requirements
Use calibration gases @ 10 PSIG, 3.0 SCFH (1.4 L/min.)
O2 Span Gas:
Air or from 1.0 to 100% O2, balance N2
Comb Span Gas:
60 to 80% (ppm ranges) of the selected comb. recorder output range
in certified mixtures of
CO + H2, 3-4% O2, balance N2.
O2 Zero Gas &
Comb Zero Gas:
System Compliance
EMC Compliance:
Safety Compliance:
2% or from 0.1 to 10% O2, balance N2
2004/108/EC
73/23/EEC
NOTES:
1.
All static performance characteristics are with operating variables constant.
2.
System accuracy referenced to 0.1 to 10% calibrated range.
3.
Response is to calibration gas (without flame arrestors).
1-2 | Thermox WDG-V / VC / VCM
Sensor : WDG-VCM
Output Range
Oxygen
Combustibles
Methane
From 0-1% to 0-100%
From 0-1% to 0-100%
From 0-2,000 ppm to 0-10,000 ppm
0 to 5%
Accuracy
Oxygen
± 0.75% of measured value or ± 0.05% oxygen, whichever is greater
Combustibles
± 2% of full scale output range
Methane
±5% of full scale output range
Response O2
90% of a process step change < 6 secs.
Drift
< 0.1% of cell output per month (< 0.005% O2 per month with 2% O2
applied)
Aspirator Air
Requirements
≈ 3 scfh (1.4 L/min) at 15 to 100 psig (1.05 to 7.04 kg/cm2)
Max. Flue Gas Temp / Probe Type / Lengths
1300°F (704°C)
1875°F (1024°C)
3000°F (1649°C)
316 SS
310 SS
Hexoloy®
Max. Sample
Dewpoint
392°F (200°C)
Sample Pressure
± 6 inch water gauge
36” - 108” (91 cm - 271 cm)
36” - 108” (91 cm - 271 cm)
24” - 72” (60 cm-182 cm)
Environment
Ambient Temp.:
-13°F to 149°F (-25°C to 65°C);
-5°F to 140°F (-20°C to 60°C) with Div. 2 Option
Relative Humidity: 10% to 90%, non-condensing
Max. Altitude:
5000 meters
Enclosure
Hinged NEMA 3R, weather resistant, stainless steel. Optional hinged NEMA 4X (IP56), explosion-proof, purged, Div II.
Power Requirements
115 VAC, ± 10%, 47-63 Hz, 700VA max.
(750 VA max. floor mount option)
230 VAC, ± 10%, 47-63 Hz, 1950 VA max.
(2000 VA max. floor mount option)
Specifications | 1-3
Calibration Gas Requirements for Combustibles Detector
Use calibration gases @ 10 PSIG, 3.0 SCFH (0.70 kg/cm2, 0.7 L/min.)
O2 Span Gas:
O2 Comb &
CH4 Zero Gases:
Comb Span Gas:
CH4 Span Gas
System Compliance
EMC Compliance:
Safety Compliance:
Air or from 1.0 to 100% O2, balance N2
From 0.1 to 10% O2, balance N2
60 to 80% (ppm ranges) or 40 to 60% (% ranges) of the selected
comb. recorder output range in certified mixtures of
CO + H2, 3-4% O2, balance N2.
2% CH4, 8% to 10% O2, balance N2
2004/108/EC
73/23/EEC
NOTES:
1.
2.
3.
All static performance characteristics are with operating variables constant.
System accuracy referenced to 0.1 to 10% calibrated range.
Response is to calibration gas (without flame arrestors).
1-4 | Thermox WDG-V / VC / VCM
Remote Calibration Unit (RCU)
O2 Only RCU
O2 & Comb. RCU
O2, Combustibles & Methane RCU
Enclosure
UL Type 4X (NEMA 4X (IP56), Div II
Environment
Ambient Temperature:
Humidity:
Max Altitude:
IEC Installation
IEC Pollution Degree 2
-18°C to 60°C
0 to 90%, non-condensing
2000 Meters
Category II
System Compliance
EMC Compliance: 2004/108/EC
Safety Compliance: 73/23/EEC
Specifications | 1-5
This page intentionally left blank.
1-6 | Thermox WDG-V / VC / VCM
SYSTEM OVERVIEW
Sensor Operations
Designed for fast response in a wide range of flue gas applications, the sensor
mounts directly on the combustion process to provide continuous measurement
of oxygen, or oxygen and combustibles, or oxygen, combustibles and methane.
A sample is drawn from the process stream by means of an air-operated aspirator and is immediately returned to the process. A portion of this gas rises past
the oxygen measuring cell and combustibles detector and returns to the primary
loop. All sample wetted parts are heated to 200°C maintain the sample above the
dewpoint of the gas.
Follow these cautions when working on the sensor:
Do not turn on the aspirator until the analyzer has been turned on
and is hot - preferably 24 hours. Also, if you turn off the analyzer or
the process is shut down, turn off the aspirator air to avoid plugging
problems!
Remove AC power from the sensor and allow the sensor to cool for
at least one hour before performing any maintenance or troubleshooting activities.
The outside of the sensor cover and all sensor assembly components
are hot during normal operation (up to 500°F, 260°C inside the
cover). Allow sensor components to cool for at least an hour before
working inside the sensor. Use caution and wear appropriate gloves
when handling components or when touching the sensor cover!
System Overview | 2-1
Sensor
Basic Elements of the Sensor
The WDG-V series analyzer consists of the following basic systems:
The Plumbing
All inlet and outlet tubing, O2 cell housing, and the combustibles flow block.
Also includes the calibration gas inlet and aspirator assembly, flow sensor,
and the aspirator used to pull the sample into the sensor.
The Measuring System
The oxygen cell, combustibles detector, methane detector, and flow sensor.
The Temperature System
The electrical cell heater (furnace), sensor case heaters, the type “K” thermocouple (maintains cell operating temperature). The sensing cell operates at a
constant temperature. The circuit board in the sensor electronic box switches
power to the furnace from the AC mains connected to the sensor. This board
also provides cold junction compensation to the thermocouple circuit.
The WDG-V is a complete standalone unit. No separate controller is required. All
required I/O is contained within the WDG-V including:
•
Qty (3) configurable 4-20 mA outputs
•
Qty (2) system alarm relays (Service and Data Valid)
•
Qty (3) configurable process alarm relays
•
Qty (1) RS485, 2 wire, MODBUS RTU interface (57.6 Kbaud)
•
Qty (1) Contact input (for remote auto calibration trigger)
•
Valve controls for Remote Calibration Unit (RCU)
The WDG-V is configured, calibrated and monitored via the MODBUS RTU
interface. AMETEK provides two options for communicating with the WDG-V:
•
AMEVision Display User Interface
•
WDG-V Configurator Software via PC (Provided with each unit)
Since the interface is industry standard MODBUS, custom user software can also
be written. For this purpose, consult the factory for the MODBUS register definitions.
2-2 | Thermox WDG-V / VC / VCM
The Oxygen Measuring Cell
The sensing element itself is a closed-end tube made from ceramic zirconium
oxide stabilized with an oxide of yttrium or calcium. Porous platinum coatings on
the inside and outside serve as a catalyst and as electrodes. At high temperatures
(generally above 1200°F/650°C), oxygen molecules coming in contact with the
platinum electrodes near the sensor become ionic. As long as the oxygen partial
pressures on either side of the cell are equal, the movement is random and no net
flow of ions occurs. If, however, gases having different oxygen partial pressures
are on either side of the cell, a potentiometric voltage is produced (See Figure
2-1). The magnitude of this voltage is a function of the ratio of the two oxygen
partial pressures. If the oxygen partial pressure of one gas is known, the voltage
produced by the cell indicates the oxygen content of the other gas. A reference
gas, usually air (20.9% O2), is used for one of the gases.
Figure 2-1. Zirconium oxide cell principle of operation.
Since the voltage of the cell is temperature dependent, the cell is maintained at
a constant temperature. The oxygen content is then determined from the Nernst
equation:
E=
O
RT
In 1
O2
4F
where R and F are constants, T is absolute temperature, and O1 and O2 are the
oxygen partial pressures on either side of the cell.
System Overview | 2-3
For measuring oxygen in non-combustibles gases, the calibration of an analyzer
is obtained from the formula:
E = A*T*Log
20.9%
O2 Unk%
AT = 48.0 at 680°C
Where A is a constant, T is the cell temperature on an absolute scale (°C + 273)
and O2 Unk% is the unknown oxygen concentration of the gas to be analyzed,
and which is calculated by the analyzer.
The cell produces zero voltage when the same amount of oxygen is on both sides,
and the voltage increases as the oxygen concentration of the sample decreases.
The voltage created by the difference in the sample gas and the reference air is
carried by a cable to the microprocessor control unit, where it is linearized to an
output signal.
NOTE
Because of the high operating temperature of the cell, combustible
gases that are present may burn. When this occurs, the cell will
generate high millivolts and cause the display to indicate less oxygen
than is actually in the gas (net oxygen content).
Hydrocarbons
When hydrocarbons are present in the gas sample, a combustion process occurs
when this gas sample is exposed to the high temperature of the zirconia cell. An
indication that hydrocarbons may be present in the gas sample is that the oxygen
reading will be lower than expected. For example, if a calibration gas cylinder
has an oxygen value of 20 PPM and 5 PPM of hydrogen (balance nitrogen), the
oxygen analyzer will read 17.5 PPM oxygen. The reduction of oxygen is due to
the combustion process where 5 PPM of hydrogen will combine with 2.5 PPM of
oxygen to form water. Thus, the oxygen analyzer will read 17.5 PPM rather than
the actual 20 PPM oxygen.
NOTE
2-4 | Thermox WDG-V / VC / VCM
The amount of oxygen reduction is dependent on the type of hydrocarbons present in the sample gas.
Sensor Operations - WDG-VCM
Close-Coupled Extrative Flue Gas Analyzer
The WDG-VCM analyzer is ideal for natural gas-fired power boilers or for those
using gas during start-up and shutdown. With this analyzer, you can monitor
oxygen and combustibles for maximum fuel efficiency. In addition, the methane
detector can be used for purge and lightoff cycles during start-up and shutdown.
WDG-VCM analyzers with a combustibles detector are used for processes that
use natural gas for start-up and oil, fuel gas or coal for primary fuel. Process gas
is drawn into the analyzer’s primary sample loop by an aspirator and returns to
process. Gas enters the split flow block that is separted by a baffle and contacts
the hot-wire catalytic combustibles and methane detector. It then passes the zirconium oxide cell, where the oxygen concentration is measured, before returning to
the main sample loop and then back to the process.
Detectors are very delicate and should be handled with care.
NOTE
System Overview | 2-5
Detectors
Combustibles
The Combustibles Detector is a dual element device. The elements differ only in
that one is coated with a catalyst. The catalyst causes oxidation to occur at a lower-than-normal temperature. The temperature of the catalyzed element changes
as the combustible mixture burns. The temperature change causes the resistance
of the catalyzed element to change. The resistance change is interpreted by the
microprocessor and the corresponding combustibles reading is displayed.
The Thermox catalytic combustibles detector will detect combustible gas present
in a sample. The sample must, however, contain enough oxygen to fully burn the
combustible present. The combustibles detector responds to all unburned combustibles gases. This includes gases such as CO and H2. You should attempt to
match your calibration gas with a mixture of CO and H2 that most closely matches the combustibles mix in your process. We recommend using an even mixture
of CO and H2 as the combustibles component within your span gas.
Examples:
•
Flue gas containing 1000 PPM CO, 1000 PPM H2 and 1% O2 with an H2 O,
CO2 and N2 balance will be sensed to contain 2000 PPM combustibles by the
catalytic detector (provided the combustibles calibration gas consists of equal
parts CO and H2 ).
•
Flue gas containing 1000 PPM CO, 1000 PPM H2 and 500 PPM O2 with a H2
O, CO2 and N2 balance will be sensed to contain only 1000 PPM Comb. by
the catalytic detector, since there is less than stoichiometric oxygen present.
In this case, the combustibles detector output will be set to full scale/
•
If there is little or no other combustible component in the flue gas, the combustibles sensor can be calibrated on a mixture of CO and air to give a CO
meaasurement.
2-6 | Thermox WDG-V / VC / VCM
Methane
The Methane Detector is a dual element device. The elements differ only in that
one is coated with a catalyst. The catalyst causes oxidation to occur at a lowerthan-normal temperature. The temperature of the catalyzed element changes as
the mixture burns. The temperature change causes the resistance of the catalyzed
element to change. The resistance change is interpreted by the microprocessor
and the corresponding reading is displayed.
The Thermox catalytic methane detector will detect methane gases present in
a sample. The sample must, however, contain enough oxygen to fully burn the
methane present.
The methane span gas should be 2% CH4 with 8 - 10% oxygen present in the
mixture.
Flow Sensor
Overview
The aspirator creates a vacuum and pulls the process gas into the analyzer. There
are two paths for the gas to take.
1. Immediately back to the process gas.
2. Through the sample loop, past the cell and/or combustible detector.
Calibrated at the factory, the flow sensor measures the sample flow through the
sample loop. The flow measurement is updated every 90 seconds. This delay is
used to allow the sensor to effectively auto-zero to prevent false alarms.
System Overview | 2-7
Alarm Limits
The flow sensor reports flow as relative to ideal (ie. 100% is ideal flow, 50% is
half of flow). An alarm is generated when the sample flow is less than 45% ideal.
The flow measurement is directional, with reverse flow generating an alarm.
Troubleshooting
The flow sensor contains two resistive elements. Each element should be approximately 8-10 Ω.
A Flow Sensore failed alarm will be generated if either element is
open.
NOTE
To Measure:
-
Turn off the instrument.
-
Measure the resistance of each element. (Flow Heater, Flow Sense) at the
terminals inside the electronics enclosure.
-
If resistance is much lower or higher than 8-10 Ω, the flow sensor must be
replaced.
Flow Sensor Heater Element (Black Wire)
Flow Sensor Sense Element (Brown Wire)
2-8 | Thermox WDG-V / VC / VCM
Temperature Sensors
There are two tightly controlled temperature zones in the WDG-V. The Oxygen
Cell Temperature and the combustible block temperature (CM units only). (On
Oxygen only units the box temperature is controlled with an RTD mounted on the
aspirator block.)
Cell Temperature
The Oxygen cell is controlled to 680°C. The cell temperature is measured by
dual, Type K thermocouples. Dual thermocouples are used for self check of the
cell temperature. The thermocouples are collocated in a single assembly. There is
no distinction between the two thermocouples as they are effectively at the same
location in the assembly.
Combustible/Box Temperature
The combustible block/ box temperature is controlled to 225°C. This temperature
is measured by 100 ohm platinum RTD. The RTD is embedded in the combustible block. On oxygen only units, the RTD is mounted on the aspirator block. All
internal plumbing components are maintained above 200°C.
Heaters
There are two main heaters in the WDG-V, the cell heater and box heater. In CM
units, an additional heater is used for fine temperature control of the combustible
block.
Cell Heater
The Oxygen Cell heater is a 300W coil heater encompassing the cell housing and
providing precision heat to the oxygen cell.
Box Heater
The box heater is a 350W cartridge heater located in the aluminum heater block
at the bottom of the aspirator. This heater provides the heat for the enclosure to
maintain all plumbing components above 200°C.
The combustible heater is a 50W heater used for precise temperature control of
the combustible Block.
System Overview | 2-9
Electronics
•
One assembly - mounted to a plate
•
The electronics/software completely control the sensor.
- No external controllers are required.
•
The configuration/monitoring interface to the electronics is RS485 MODBUS
RTU.
•
Sensor wiring and Customer wiring (i.e. power, I/O) are directly connected to
the main electronics.
- Connections are pluggable.
- Both plugs and headers are laser printed for easy matching.
I/O
The WDG-V has the following I/O available:
•
Qty (3) configurable 4-20 mA outputs
•
Qty (2) system alarm relays (Service and Data Valid)
•
Qty (3) configurable process alarm relays
•
Qty (1) RS485, 2 wire, MODBUS RTU interface (57.6 Kbaud)
•
Qty (1) Contact input (for remote auto calibration trigger)
•
Valve controls for Remote Calibration Unit (RCU)
2-10 | Thermox WDG-V / VC / VCM
Analog Outputs
There are 3 analog outputs in the WDG-V. Each analog output can be configured
for Mode, Function, and Span, Zero.
Mode options include, 0-20ma, 4-20ma and NAMUR.
Functions include; Oxygen, cell millivolts, combustibles*, methane*
The analog outputs can be powered internally or externally. External power must
be greater than 26V.
The NAMUR outputs are implemented as follows:
4-20 mA signal
0 mA
3.5 mA
3.8 mA
4 to 20 mA
20.5 mA
Analyzer Condition
Analyzer unpowered, or completely failed
Critical Alarm - analyzer reading unusable (factory default)
Reading Under Range (Example: user sets range to 2-10%. Current reading is 1.9%
Normal Operation
Reading Over Range (Example: range is 0-10%. Current reading is 12%
* For combustible and/or methane units only.
Alarm Contacts
There are 5 alarm contacts in the WDG-V, 1 service alarm, 1 data valid alarm,
and 3 configurable process alarms.
Service Alarm Relay
This relay is used to determine if there is a problem with the analyzer. This relay
is normally open. Contacts are closed when in normal mode. The relay will be
open when there is a problem with the analyzer or on loss of power.
Data Valid Relay
This relay is used to determine if the output is representative of the process. This
relay is normally open. Contacts are closed when in normal mode. Contacts are
open when there is a problem with the analyzer, when the analyzer is calibrating,
warming up, when the analyzer is in diagnostic mode, or on loss of power.
System Overview | 2-11
Contact Input
There is one Contact Input peripheral on the WDG-V.
An auto calibration can be initiated using the contact input peripheral. (Only with
the RCU option).
Auto calibration must be selected in the sensor configuration menu.
Remote Calibration Unit Interface
The WDG-V has 5 solenoid valve drivers that are compatible with AMETEK’s
Remote Calibration Unit (RCU).
See wiring section for interface details.
Communications Interface
There is one communications interface on the WDG-V.
•
RS485 MODBUS RTU
Configuration:
- Two wire
-57.6 Kbaud
- No parity
- 1 stop bit
The WDG-V is configured, calibrated and monitored via the MODBUS RTU
interface. AMETEK provides two options for communicating with the WDG-V:
•
AMEVision Display User Interface (Reference separate manual)
•
WDG-V Configurator Software via PC (Provided with each unit)
Since the interface is industry standard MODBUS, custom user software can
also be written. For this purpose, refer to Chapter 11 for the MODBUS MAP or
consult the factory for the MODBUS register definitions.
2-12 | Thermox WDG-V / VC / VCM
Flame Arrestors
The WDG-V sensor contains two flame arrestors to prevent it from being an ignition source to the process for short periods of high combustible levels (25% of
LEL - Lower Explosive Limit) in the process. The flame arrestors are not, however, intended to protect the process where the combustibles’ levels are constantly
high.
System Overview | 2-13
Start-Up Checklist
Review the installation chapter in the User’s Manual.
Install the sensor on the process.
-
If the sensor is on the pipe nipple, insulate the pipe from the process wall
to the mounting flange on the sensor.
-
Install the probe.
-
Install the probe heater, if required.
This type of sensor requires a clean, dry instrument air supply for the aspirator. Connect the air supply to the aspirator air supply inlet but DO NOT turn
on the air until the sensor is at operating temperature. If the analyzer is turned
off or the process is shut down, the aspirator air should be turned off to avoid
plugging problems.
If used, interconnect the AMEVision Display User Interface and the sensor
per the appropriate interconnecting drawing.
-
The specified cable must be used in order to ensure proper operation of the
systems.
Connect AC mains supply to the sensor. Make sure that the AC voltage is
within specified limits.
2-14 | Thermox WDG-V / VC / VCM
Common Operator Errors
If you follow the steps below, your analyzer will operate with a minimum of
maintenance and troubleshooting.
Connect the air supply to the aspirator air supply inlet but DO NOT turn on
the air until the sensor is at operating temperature. If the analyzer is turned
off or the process is shut down, the aspirator air should be turned off to avoid
plugging problems.
If your process is running and the analyzer is installed on the process, the
analyzer must have power applied to it to prevent plugged plumbing and sensor component damage. The case heaters must be on if the process is running
to maintain sensor plumbing above 200°C.
Do not use pipe dope or any other contaminant that gives off combustible
vapor on any joints of the sample tubing.
For O2 calibration gases, do not use calibration gases to check Thermox analyzers if they contain a mixture of oxygen and combustibles. Note, however,
that this is acceptable for combustibles span gases (used only if you purchased the combustibles option).
Always introduce calibration gases at the recommended flow rate of 3.0 scfh.
When working on the plumbing inside the sensor cabinet, turn the power off.
Use caution if performing maintenance on the sensor while the process is
running. Removing any part of the sensor can allow process gases and gases
of high temperature to escape into the sensor.
Do not handle the cell excessively. Do not try to clean the cell except by
rinsing. If you need to handle the cell, grasp by touching the seal fitting at the
top; never touch the bare part of the cell.
Do not remove a cell or type “K” thermocouple that you may want to use
again when the inside of the furnace is still hot - severe thermal shock can be
destructive to either of them.
Always replace the metal cell O-ring when replacing the oxygen cell.
System Overview | 2-15
Technical Support
AMETEK/Thermox is committed to providing the best technical support in the
industry. If you need service or application assistance, please call AMETEK at
(412) 828-9040, or your local AMETEK/Thermox representative.
Before you call the factory for technical support, run test gases and record the
following values (you may be asked by the factory to provide this information
when receiving service):
Cell millivolts
Thermocouple millivolts
Cell temperature
If you need to return equipment, you will be asked to provide the following information before obtaining a Return Material Authorization (RMA) number.
Billing and shipping address
Model number
Serial number
Purchase order number
Telephone number
NOTE
2-16 | Thermox WDG-V / VC / VCM
Before returning material, you must get an RMA number from the
factory.
INSTALLATION GUIDE
NOTE
The operations in this chapter should be performed only by qualified
service personnel experienced in electrical safety techniques. Never
service the sensor unless power has been removed from the ensor,
and the sensor has been allowed to cool for at least one hour. Also,
always use gloves when working on the sensor.
Do not turn on the aspirator until the sensor has been turned on
and is hot - preferably 24 hours. If you turn off the analyzer or the
process is shut down, turn off the aspirator air to avoid plugging
problems!
Installation Guide | 3-1
Analyzer Location
Observe the following guidelines when selecting an analyzer installation location:
1. Select a readily accessible position for the analyzer to allow for routine maintenance. Comfort levels for maintenance personnel should be considered in
placement of the sensor and AMEVision Display User Interface.
2. The installation location should be free from excessive vibration and the
ambient temperature is required to be within the limits listed in the specifications. If the ambient temperature is outside the specified limits or the vibration is excessive, please contact Thermox Sales or Service Department at
(412) 828-9040. We have special options to address ambient temperatures
outside the listed specifications.
This chapter shows you how to install your WDG-V sensor components and
AMEVision Display User Interface, and includes the following sections:
•
•
•
Inspection of Shipping Contents
Mechanical Installation
Wiring
If you have an oxygen-only analyzer without the combustibles option, the
combustibles and methane options described in this chapter do not apply. These
combustibles and methane sections are clearly identified. Likewise, if you have
the combustibles option, but not the methane option, you can skip all sections
pertaining to methane.
Inspect Shipping Contents
Remove any packing material from the WDG-V sensor. Check for damage. If any
is found, notify the shipper.
3-2 | Thermox WDG-V / VC / VCM
Mechanical Installation
This section describes how to mount and connect gases to your WDG-V sensor.
This includes the following:
•
Sample Inlet Probe Installation
•
Probe Heater Installation (optional)
•
Exhaust Tube Installation (optional)
•
Sensor Mounting
•
Remote Calibration Unit Mounting and Plumbing (optional)
Sample Inlet Probe Installation
NOTE
For some applications, you must first connect the sample inlet probe
to the sensor before mounting the sensor. The same applies if you are
installing the probe heater or exhaust tube. Therefore, these subjects
are discussed before the sections on how to mount the sensor to the
process.
The sample inlet probe is how the sample gas gets to the sensor. Aspirator air
within the sensor is used to pull process gas from the sample probe into the sensor. The sample inlet probe is identified by the threads on both ends.
•
Figure 3-1 shows how to install the standard 1/8” NPT probe.
•
Figure 3-2 shows how to install a ceramic or long probe.
•
Figure 3-3 shows port locations on the rear of the sensor.
You can connect the probe to the sensor before mounting the sensor unless the
probe is either very long or made of ceramic. In that case you should mount the
sensor first and then install the probe.
Installation Guide | 3-3
Standard 1/8” NPT probe installation
Connect a standard 1/8” NPT sample inlet probe to the sensor as follows:
•
If your probe is 4 feet or less, simply screw the probe into the back of the
sample probe port on the rear of the sensor, then mount the sensor - see Figure 3-1.
Standard Probe lengths
3', 4', 5', 6' and 8'
Threaded probes screw
into fitting
Exhaust extension
3/8"T x 10-3/8"Lg
(optional). Threaded
extension screws
into fitting.
Backplate
Bushing
3/4"NPT(F) x 3/8"NPT(F)
(NA for long probes)
Insert Plate
Fitting
3/8"P x 3/8"T
tapped 1/8"NPT
(NA for long probes)
Probe Inlet Hook
TOP VIEW
Probe Support
54-1/8"long
Assembly replaces
bushing for probes
5', 6 and 8'
Figure 3-1. WDG-V Probe installation.
3-4 | Thermox WDG-V / VC / VCM
Ceramic probe or long probe installation
If you ordered a long probe or a ceramic probe, connect the probe to the sensor as
follows:
1. Mount the sensor. See the “Sensor Mounting” section later in this chapter
for help. Note, however, that if you are installing the probe heater or exhaust
tube, you must connect these items to the sensor before mounting the sensor.
2. Remove the probe inlet hook from the sensor as shown in Figure 3-2. To do
this, loosen the fitting connecting the inlet hook to the aspirator block and
pull the hook out of the sensor.
3. Unscrew the bushing that was plugged into the sensor inlet port and discard.
The probe comes with its own bushing and fitting.
4. Thread the sensor probe into the sensor inlet port on the sensor backplate. Do
not let the probe hit against the Thermox-provided pipe nipple as you insert
the probe into the process. If inserting a ceramic probe, you must slowly
insert it into a hot process to prevent the probe from cracking due to thermal
shock.
5. Reconnect the probe inlet hook to both the sensor backplate and the aspirator
block and tighten.
Backplate
Bushing
Aspirator Block
Probe Inlet Hook
TOP VIEW
Figure 3-2. WDG-V ceramic probe installation - top view.
Installation Guide | 3-5
Probe Heater Installation
Do not install the probe heater if your sample gas contains a potentially explosive mixture of combustibles; the probe heater can heat
the flue gas to the point of ignition.
The probe heater must be connected to the sensor before mounting
the sensor to the process.
NOTE
For installations where a standard setup allows the sample gas to cool below its
acid dewpoint before reaching the sensor, you can attach the optional probe heater to the probe heater port on the backplate of the sensor to heat the sample gas
- see Figure 3-3. This will prevent plugged plumbing due to condensation of the
sample gas. For example, when a fuel containing only a few thousand parts per
million of sulfur is burned with a high level of excess oxygen (over 4 percent), the
resulting sulfuric acid mist may condense at temperatures as high as 175°C.
The probe heater comes shipped in an envelope, is 8.5” long, has a 3/8 NPT
bushing and wires leading out of it.
Sample Probe
Inlet Port
Exhaust Port
Probe Heater Port
(Remove Plug)
Terminal Box
Figure 3-3. Port locations on rear of sensor - front view.
3-6 | Thermox WDG-V / VC / VCM
Follow the steps below to install the probe heater:
1. Remove and discard the plug from the probe heater port as shown in Figure
3-3. This port is plugged during shipment and should remain plugged if no
probe heater is installed.
2. Screw the probe heater into the 3/8” NPT probe heater port.
3. Route the wires from the probe heater to the electronics enclosure.
4. Remove the electronics box cover and connect the wires to an AC power
source.
Exhaust Tube Installation
If installing an exhaust tube, you must connect it to the sensor before
mounting the sensor to the process.
NOTE
An optional stainless steel exhaust tube is available (PN 70619KE), which allows
you to extend the aspirator exhaust.
To install the exhaust tube, thread the exhaust tube into the exhaust port on the
rear of the sensor as shown in Figure 3-3.
Sensor Mounting
•
The sensor ambient temperature range is -25°C to 65°C.
•
For purged units, the sensor ambient temperature range is -20°C to 60°C.
•
When installing a probe heater or exhaust tube along with the sensor, attach it
to the back of the sensor before mounting the sensor to the process.
Do not open a Z-purge, hinged-sensor enclosure door until you first
verify that the area has been classified as nonhazardous.
Installation Guide | 3-7
Methods for mounting the sensor
There are three methods for mounting a WDG-V sensor to the process:
•
Flush with the process,
•
Using a customer-supplied pipe nipple,
•
Using a customer-supplied flange.
Each of these methods is explained in the sections that follow. The flush mount
method is preferred because it positions the sensor as close to the process as possible and minimizes the chances of the sample gas cooling below its dewpoint.
Mounting dimensions for the hinged style enclosure, including the Z-purge option, are provided in Figure 3-4.
NOTE
3-8 | Thermox WDG-V / VC / VCM
If the protection nipple provided by AMETEK/Thermox extends past
the inside of the refractory wall on a process that has a gas temperature over 1200°F (650°C), the nipple should be cut down before
installation in order to remain flush with the inside wall.
8.48
215.50
4.39
111.44
Exhaust
Restrictor
Pressure
Gauge
`
3.80
96.59
18.61
472.59
Hinge
Side
2.51
63.86
Instrument
Air Inlet
5.23
132.95
Metering
Valve
Probe CL
3.64
92.56
Check
Valve
Terminal Box
Figure 3-4. Mounting dimensions - Hinged sensor.
A 7/16” wrench is required to open both doors to the sensor.
NOTE
Installation Guide | 3-9
Sensor flush mount (Preferred Method)
Weld the 7” x 7 5/8” (17.78 cm x 19.37 cm) mounting plate (P/N 70626SE) supplied with the sensor to the process wall over a 3 3/4” (9.52 cm) diameter hole
as shown in Figure 3-5. Then bolt the sensor to the mounting plate. This method
positions the sensor as close to the process as possible and minimizes any chance
that the sample gas will cool below its dewpoint.
Sensor bolted to mounting plate after
plate is welded to wall by customer.
Gasket (by Ametek)
Mounting plate supplied by Ametek
with 2 7/8"OD x 9"long nipple (7.30cm
x 22.86cm) welded to wall over 3 3/4"od
(9.52cm) hole by customer.
Process wall
Probe
Single stud at
12 o'clock position
SIDE VIEW
Proper orientation for
mounting plate to flange
Figure 3-5. WDG-V sensor flush mount (preferred).
Customer-supplied pipe nipple sensor mounting
If you can’t mount the sensor flush with the process as described in the previous
section (for example, the process wall is curved sharply, covered with insulation,
or obstructed with reinforcing members) you can instead:
1. Weld a short 3 1/2” diameter pipe nipple to the process wall and make sure to
extend the nipple through the skin of the process (see Figure 3-6).
3-10 | Thermox WDG-V / VC / VCM
Sensor bolted to mounting plate after
plate is welded to nipple by customer.
Gasket (By Ametek)
Mounting plate (supplied by Ametek)
1 7/8"OD with 9"long nipple (7.30cm
x 22.89cm) welded over 3 3/4"OD
(9.52cm) hole by customer.
Process wall
Probe
Required 1" (25.4cm) minimum
insulation around exposed portion
of nipple
Pipe nipple
(supplied by customer)
SIDE VIEW
Single stud at
12 o'clock position
Proper orientation for
mounting plate and flange
Figure 3-6. Customer-supplied pipe nipple sensor mounting.
2. Weld the 7” x 7 5/8” (17.78 cm x 19.37 cm) supplied plate to the end of the
nipple, taking care to center it over the hole
3. Wrap the nipple with at least 1” (2.54 cm) weatherproof insulation and heat
trace it.
4.
Then bolt the sensor to the plate.
If mounting the sensor using this method, you can also install the probe heater to
minimize the possibility of the sample gas cooling below its dewpoint.
Installation Guide | 3-11
Flange sensor mounting method
You can also mount the sensor to a 3” x 300# customer-supplied flange. In this
case, you will not need the AMETEK-supplied mounting plate with attached
protective pipe
nipple. If you have a different type of flange,
AMETEK supplies a
1
2
number of flange adapters.
REV.
A
B
REVISION DESCRIPTION
DATE
ORIGINAL ISSUE
ADD MISSING PARTS
06-11-13
10-14-13
DR
APPR
SB
SB
MG CAM
CUSTOMERS 3"-300# FLANGE (REDUCING
FLANGE IF REQUIRED) CONNECTED TO
2 1/2" MINIMUM SCH 40 PIPE ATTACHED
TO WALL BY CUSTOMER.
B
B
PROCESS WALL
GASKET BY THERMOX
1" MIN INSULATION AROUND
EXPOSED PORTION OF NIPPLE
3"-300# FLANGE MOUNTING
A
A
PRODUCT LINE
THERMOX
DIMENSIONS ARE IN INCHES
TOLERANCES:
FRACTIONAL 1/32"
ANGULAR: MACH 30'
TWO PLACE DECIMAL
.010
THREE PLACE DECIMAL .005
SURFACE ROUGHNESS 125
CONFIDENTIAL
REPRODUCTION PROHIBITED WITHOUT
PERMISSION BY AMETEK P & AI DIVISION
Figure 3-7. WDG-V sensor flange mount.
MATERIAL
2
3-12 | Thermox WDG-V / VC / VCM
SEE BILL OF
MATERIALS
SB
06-11-13
DOC. CONT.
MG
10-8-13
ENGR. APPR.
CAM
10-8-13
ORIG. ISSUE
CK
DRAWN BY
SCALE
NTS
WDG-V PURGE SENSOR
MOUNTING & DIMENSIONS
SIZE
06-11-13
SHEET
1
OF
AMETEK
PROCESS & ANALYTICAL INSTRUMENTS DIV.
150 FREEPORT ROAD
PROCESS & ANALYTICAL INSTRUMENTS DIVISION PITTSBURGH, PENNSYLVANIA 15235 USA
1
A
DWG. NO.
202-645-3
1
REV
B
Calibration/Aspirator Air
Required calibration gases and tubing
See Chapter 2 for information on calibration gases. The span gas is
the high gas; the zero gas the low gas.
NOTE
The span gas must be 10 times greater than the zero gas. For example, if the zero
gas is 1% O2, the span gas must be 10% O2 or higher.
O2 Span Gas
Instrument air (20.9%) or from 1.0% to 100% O2.
O2 Zero Gas
From .1 to 10% O2, balance N2
Combustibles Span Gas
If using a WDG-VC analyzer, you must also have a combustibles span gas as
follows:
•
60 to 80% of the full combustibles operating range in certified mixtures of
CO, with an equal amount of O2, balance N2.
For example, if using an operating range of 0 to 2000 ppm combustibles, use
1200 ppm CO, 1 to 2% O2, balance N2.
Methane Span Gas
If using a WDG-VCM analyzer, you also must have a methane span gas as
follows:
•
2% CH4, 8 to 10% O2, balance N2.
Installation Guide | 3-13
Z-Purge Connections (Optional)
Connect instrument air to the sensor Z-purge inlet. Maintain the pressure and
flow as directed on the Z-purge warning label.
•
You must provide a protective gas supply with an alarm for a loss of pressure
indication in order to meet regulatory requirements.
•
You must not exceed the maximum temperature rating listed in the specifications.
•
Process pressure is assumed to always be subatmospheric (vacuum) unless
the High-pressure option is installed.
•
The protective gas supply shall be equipped with an alarm to indicate failure
of the protective gas supply to maintain the minimum pressurized enclosure
pressure.
•
The indicator shall be located for convenient viewing.
•
The indicator shall indicate the enclosure pressure.
•
The sensing point for the indicator shall be located to take into account the
most onerous conditions of service.
•
There shall be no devices between the pressurized enclosure and the protective gas supply alarm other than an isolating valve and/or a pressure of flow
controlling mechanism.
•
Any isolating valve shall
- Be marked
WARNING - PROTECTIVE GAS SUPPLY VALVE
FOLLOW INSTRUCTIONS BEFORE CLOSING
- Be capable of being sealed or secured in the open position.
- Have an indication of whether it is open or closed.
- Be located immediately adjacent to the pressurized enclosure.
- Be used only during servicing of the pressurized enclosure.
•
Any pressure or flow controlling mechanism, if adjustable, shall require a
tool to operate it.
•
No filters shall be fitted between the pressurized enclosure and the protective
gas system alarm.
•
No isolating valve shall be fitted between the indicator and the pressurized
enclosure.
3-14 | Thermox WDG-V / VC / VCM
•
The WDG-V sensor contains two flame arrestors to prevent it from being an
ignition source to the process for short periods of high combustible levels
(25% of LEL - lower explosive limit) in the process. The flame arrestors are
not, however, intended to protect the process where the combustibles levels
are always high.
Do not install the probe heater if your sample gas contains a potentially explosive mixture of combustibles; the probe heater can heat
the flue gas to the point of ignition.
Z-Purge Shutdown Procedure
Hot internal parts are above the ignition temperature of combustible
gases. Power must be disconnected from the sensor for 90 minutes
while maintaining purge air flow before the door is opened, unless
the area is demonstrated to be nonhazardous. The Z-purge apparatus
provided by Thermox meets NFPA 496.
Power must not be restored after the enclosure has been opened until
the enclosure has been purged for 65 minutes at a pressure of 0.4” of
water.
Always use tubing that is free of oil and dirt. Stainless steel tubing is recommended.
Figure 3-8. Z-purge setup.
Installation Guide | 3-15
Calibration Gas Inlet Port
1/4"T Compression Fitting
Instrument
Air Inlet
1/4"NPT (F)
Purge Air Inlet
1/4"T Compression
Fitting
Figure 3-9. Sensor inlet connections.
3-16 | Thermox WDG-V / VC / VCM
Manual calibration and aspirator air connections
NOTE
If you have a remote calibration unit (RCU), skip the remainder of
this section and proceed to the “RCU Mechanical Installation” section.
Aspirator air connection
If you have an AMEVision Display User Interface, you can use the
flow gauge to set the flow as described in Chapter 5.
NOTE
•
Connect aspirator air to the aspirator air inlet on the sensor (see Figure 3-9).
The recommended initial aspirator pressure setting is 3 psi. This will vary
depending on the pressure/vacuum of the process for which the WDG-V is
installed. If you purchased an AMEVision Display User Interface, you can
use the Flow Sensor output diagnostic screen to set the flow to the ideal rate.
If you do not have an AMEVision Display User Interface, it is recommended
to adjust the aspirator until you get a response from the analyzer, then increase the pressure by 0.25 to 0.5 psi. Do not turn on the aspirator until the
sensor has been turned on and is hot (preferably 24 hours from a cold start; 1
hour after a restart). See page 5-1 for furter detail.
Calibration gas connection
•
A calibration gas inlet port is provided to allow you to calibrate the system as
shown in Figure 3-9. During calibration, you turn off the aspirator air supply and inject your calibration gases into the calibration gas inlet port on the
sensor. During normal operations, the calibration gas inlet port should be
plugged.
Installation Guide | 3-17
Remote Calibration Unit Mechanical Installation
Remote calibration unit (RCU) mechanical installation includes the following:
•
Mounting the RCU.
•
Plumbing calibration gases to the RCU.
•
Plumbing the RCU to the sensor.
This section is divided into instructions for standard RCU and combustibles RCU
mechanical installation. Go to the section that applies to the type of RCU you
purchased.
Do not turn on the aspirator until the sensor has been turned on and
is hot - preferably 24 hours. Also, if you turn off the analyzer or the
process is shut down, turn off the aspirator air to avoid plugging
problems!
Standard RCU mechanical installation
Figure 3-10 shows mounting dimensions for a standard RCU. Use two #10
screws to mount the RCU. Note from Figure 3-10 that only the top left and lower
right holes are used; the other holes hold the RCU to its mounting plate. The
ambient temperature range for the RCU is -20° to 70°C. Mount the RCU as close
to the sensor as possible. Shorter calibration plumbing improves response times,
reduces calibration gas expense, and reduces the chance of contaminants in the
calibration gas plumbing.
3-18 | Thermox WDG-V / VC / VCM
Figure 3-10. RCU mounting dimensions - O2-only.
Installation Guide | 3-19
Standard RCU
Make the following calibration gas connections to the standard RCU. Figure 3-10
shows standard RCU calibration gas connections:
1. Connect instrument grade air that is free of oil or dirt to the instrument air
inlet connection on the left side of the RCU. If you are also using instrument
air as your span gas, you can skip Step 2 below because instrument air will
be used for both aspirator air and the span calibration gas.
2. If you wish to use a span gas other than instrument air, you must purchase the
optional alternate span gas kit. Connect the span gas to the alternate span gas
inlet on the RCU if your span gas is other than instrument air.
3. Connect the O2 zero calibration gas to the O2 zero gas inlet connection on the
RCU.
4. Connect the calibration gas outlet on the right side of the RCU to the calibration gas inlet on the sensor -see Figure 3-9 for the calibration gas inlet
connection on the WDG-V sensor. Also, be sure to install the supplied check
valve as close to the sensor as possible - note the orientation of the valve.
5. Connect the aspirator air outlet on the RCU to the aspirator air inlet connection on the sensor.
Combustibles RCU
Figure 3-11 shows mounting dimensions for the combustibles RCU. Mount the
RCU as close to the sensor as possible. Shorter calibration plumbing improves
response times, reduces calibration gas expense, and reduces the chance of contaminants in the calibration gas plumbing. The ambient temperature range for the
combustibles RCU is -20° to 70°C.
Make the following calibration gas connections to the combustibles RCU.
3-20 | Thermox WDG-V / VC / VCM
COMBUSTIBLES (SPAN 2)
CALIBRATION GAS INLET
1/4” COMPRESSION FITTING
Figure 3-11. RCU mounting dimensions - Combustibles.
1. Connect the O2 span gas, if other than instrument air, to the alternate O2 span
gas inlet.
2. Connect the O2 zero calibration gas to the O2 zero gas inlet connection on the
RCU.
3. Connect the combustibles span calibration gas to the combustibles span gas
inlet on the RCU. For a WDG-VCM analyzer with RTD-type combustibles
detector, the combustibles span calibration gas would also include the methane span calibration concentration.
4. Using appropriate tubing, connect the calibration gas outlet on the right side
of the RCU to the calibration gas inlet on the sensor - see Figure 3-8 for sensor calibration gas inlet connection.
5. Using appropriate tubing, connect the aspirator air outlet on the right side of
the RCU to the aspirator air inlet connection on the sensor.
Installation Guide | 3-21
Figure 3-12. RCU flow diagram - Single Span.
Figure 3-13. RCU flow diagram - 2 Span - Combustibles.
3-22 | Thermox WDG-V / VC / VCM
Figure 3-14. RCU plumbing - 3 Span - Combustibles & Methane.
Installation Guide | 3-23
Wiring
Remove AC mains power from the controller before performing wiring.
Connections to the sensor unit are made through the wiring card. The wiring card
is located on the front-bottom of the sensor unit.
Any screw terminals on the wiring card not described in this section
are reserved for future use.
NOTE
This wiring section shows you how to make the following connections:
•
AC mains supply wiring to WDG-V sensor
•
WDG-V unit to remote calibration unit
•
WDG-V unit to alarm devices
•
WDG-V unit to current output devices
•
WDG-V unit to AMEVision Display User Interface (RS-485 communications)
•
AC mains supply wiring to AMEVision Display User Interface
In addition, this wiring section provides mandatory EMC grounding, shielding,
and noise protection requirements.
3-24 | Thermox WDG-V / VC / VCM
General Wiring and Conduit Requirements
This section describes general wiring and conduit requirements.
•
Sensor wiring conductors must be rated at a minimum of 80°C. All other
wiring conductor ratings should be for the minimum temperature required for
the equipment being connected to the analyzer, but not less than 60°C.
•
Use only the applicable NEMA-approved conduit fittings or cable fittings to
maintain the NEMA rating for the sensor enclosure or WDG-V. If not using
a conduit entry, leave the factory NEMA-approved plugs intact. Never leave
any holes unplugged.
•
Follow all applicable electrical codes for your location.
•
Follow proper grounding, shielding and noise protection practices as described in this section.
•
For all sensor and signal wiring use twisted-pair cable, 18 to 22 AWG (0.82
mm2 to 0.33 mm2 ), with an overall braided shield, or twisted-pair cable in
rigid metal conduit.
•
For AC mains supply wiring, use between 12 and 14 American Wire Gauge
(AWG) or equivalent metric between 3.3 mm 2 and 2.1 mm2.
•
Use the conduit entry point closest to the connections you are making. Do not
add any additional conduit entry holes!
SS External
Ground Stud
M5 Ring Terminal
12-10AWG
(2) 1/2"NPT
Conduit Entries
Supplied with
Nema Conduit Plugs
42464JE
(4) 3/4"NPT
Conduit Entries
Supplied with
Nema Conduit Plugs
42454JE
Installation Guide | 3-25
WDG-V Mains Supply Connections
Do not run control unit AC mains supply wiring in the same conduit
with other AC line power wires. By keeping this wiring separated,
you prevent transient signals from reaching the control unit.
The WDG-V can operate using between 115 VAC, ± 10% or 230 VAC, ± 10%. There
is no power switch or circuit breaker on the sensor, and it must be protected by
installing it on a circuit-protected line, maximum 15 amperes, with a switch or
circuit breaker in close proximity to the control unit and within easy reach of
an operator. Mark the switch or circuit breaker as the control unit disconnecting
device.
Mains supply connections to the control unit are as follows:
L - Line connection
N - Neutral connection (USA)
Chassis Stud - Equipment ground (protective conductor)
Use the 1/2” conduit entry hole in the WDG-V for AC mains supply wiring. Use
the chassis stud next to the 1/2” conduit entry hole for equipment ground (protective conductor).
AC (L) and (N) markings are provided by the terminal block for connection of
AC power. These markings are for reference purposes only, such as for use on
system wiring diagrams, etc. The system/product has or needs no specific LINE
or NEUTRAL connection for any function, safety or otherwise. The (N) terminal
is not internally grounded, nor needs to be. The system will operate normally regardless of what AC input terminal (L or N) the AC Line or Neutral is connected
to, or, if there is a Neutral used at all (i.e., 208 VAC US power connection).
3-26 | Thermox WDG-V / VC / VCM
EMC grounding, shielding, and noise protection
For EMC purposes, under no circumstances should you leave cable
shields disconnected at one end or both ends of the cable (sensor or
control unit or other device).
You must use twisted-pair cable in rigid metal conduit or use twisted pair cable
with an overall braided shield. All cable shields or conduits connecting to the
sensor unit must be chassis grounded.
SHIELD RING METHOD
Shield Terminal Ring
Quick Disconnect (1/4”)
Cable Shield(s)
Max. Length 1 inch
Capacitor
Conduit Housing
Conduit Nut
GROUND STUD METHOD
Conduit Housing
Cable Shield(s)
Max. Length 2 inches
Capacitor
Conduit Nut
Nut
Washer
Ring Terminal
Figure 3-25. Direct shield grounding methods
Installation Guide | 3-27
Transient and RFI interference
This section describes transient and RFI interference precautions:
•
Although there are transient and noise protectors on all sensor unit I/O connections (communications, current outputs, sensor, etc.), this protection is
intended to act as a last line of defense against unwanted transient and RFI
interference.
•
Proper installation practices to prevent the introduction of transients and
noise into the system must be followed. Inductive loads connected to the
sensor unit must have transient suppressors installed at the inductive loads.
Be sure to place the transient suppressor as close to the load as possible.
Examples of transient suppressors include MOVs, TRANSORBs, and RC
snubbers.
•
AC mains supply wiring should not be run in the same conduit with mains
supply wiring that feeds heavy inductive loads.
•
Avoid running signal wiring in the same cable or conduit with wires that
power inductive loads unless all the cables within the conduit are shielded,
the inductive loads are small, and transient suppressors are used at the loads.
•
Do not run signal lines in the same cable or conduit with high voltage lines.
•
For optimum noise protection, AMEVision Display User Interface mains
supply wiring should be connected to a circuit separate from any circuit that
could introduce transients into the system. As an example, do not run motors,
blowers, or air conditioners using the same mains supply circuit or conduit as
the sensor unit’s mains supply circuit or conduit.
3-28 | Thermox WDG-V / VC / VCM
Current Output Connections
Standard current outputs
There are three current outputs on the WDG-V sensor. The current output connections are labeled as follows on the sensor board terminals:
Analog Output #1 => +I1Analog Output #2 => +I2Analog Output #3 => +I3-
The current outputs are referenced as Analog Outputs 1, 2 and 3 in the AMEVision Display User Interface and PC software. Be sure to observe the polarity
when connecting current output devices to these terminals.
Each of the current outputs are capable of driving up to 1000 ohm loads.
The current outputs can be selected for the following ranges:
0-20mA
4-20mA
NAMUR
The NAMUR outputs are implemented as follows:
4-20 mA signal
0 mA
3.5 mA
3.8 mA
4 to 20 mA
20.5 mA
Analyzer Condition
Analyzer unpowered, or completely failed
Critical Alarm - analyzer reading unusable (factory default)
Reading Under Range (Example: user sets range to 2-10%. Current reading is 1.9%
Normal Operation
Reading Over Range (Example: range is 0-10%. Current reading is 12%
Installation Guide | 3-29
Figure 3-26. Wiring
3-30 | Thermox WDG-V / VC / VCM
Alarm Contact Connections
This section describes how to make wiring connections for any alarm devices
you wish to connect to the control unit.
Standard alarm connections
The WDG-V sensor provides five sets of normally open alarm contacts as follows:
Terminal Block ID
Description
SVC - Service Alarm
DV - Data Valid Alarm
ALM1
-
Process Alarm 1 (Configurable)
ALM2
-
Process Alarm 2 (Configurable)
ALM3
-
Process Alarm 3 (Configurable)
The process alarms are referenced as Relay 3, Relay 4 and Relay 5 in the
AMEVision.
Relay 3
Relay 4
Relay 5
-
-
-
ALM 1
ALM 2
ALM 3
Service Alarm - The service alarm indicates that a critical alarm exists and the
analyzer needs repair. Readings are not valid when this alarm is active. The relay
contacts are normally-open and are OPEN (de-energized) when a critical alarm
exists.
Data Valid - The data valid alarm indicates if the concentration readings are representative of the process. When this alarm is active, concentration readings may
not be representative of the process and should not be used for control/safety.
Examples of this alarm include: Warm-up mode, in calibration, and diagnostics
mode. This alarm differs from the service alarm in that there is not necessarily
a problem with the analyzer when this alarm is active. The relay contacts are
normally-open and are OPEN (de-energized) when the concentration readings are
not valid.
Installation Guide | 3-31
Remote Calibration Unit Connections
If you don’t have the remote calibration unit (RCU) option, skip this section.
Oxygen-only remote calibration unit connections
Oxygen-Only Remote Calibration Unit (RCU) connections on the WDG-V wiring card, and their RCU connections, are as follows (see Figure 3-27):
ZERO GAS - WDG-V Terminal Z to Pin 13 on RCU
ASPIRATOR - WDG-V Terminal A to Pin 14 on RCU
O2 SPAN - WDG-V Terminal S1 to Pin 15 on RCU
VALVE COMMON - WDG-V Terminal ‘V’ to pin 16 on RCU
Combustibles remote calibration unit connections
If you have the combustibles option (and a combustibles RCU), you must also
make the following connection (see Figure 3-27):
COMBUSTIBLE SPAN - WDG-V Terminal S2 to Pin 17 on RCU
For a WDG-VCM with hot-wire detector, Remote Calibration Unit connects:
METHANE SPAN - WDG-V Terminal S3 to Pin 18 on RCU
3-32 | Thermox WDG-V / VC / VCM
SPAN 3
SPAN 4
SPAN 2
SPAN 1
Z
A
VALVE COM
CIN
VALVE COM
Remote cal
switch normally
open (dry
contact closure)
Overall braided shield
or rigid metal conduit
All common wires
to pin 16 on RCU
and valve commons
on control unit
Figure 3-27. RCU connections.
Digital input to initiate remote calibration unit
The digital input connections on the wiring card allow you to initiate a remote
calibration from a remote location. For this option to work, you must have a remote calibration unit. The system monitors the digital input, and when the switch
closes, it sends the system into an automatic calibration. The switch you connect
to this digital input must be a normally open switch.
Digital input connections are labeled as follows on the wiring card:
CIN
Installation Guide | 3-33
Notes About Ducting
Ducting Between Pressurized Enclosure and Inlet
The intake ducting to a compressor should not normally pass through a hazardous area. If the compressor intake line passes through a hazardous area, it should
be constructed of non-combustible material and protected against mechanical
damage and corrosion. Adequate precautions should be taken to ensure that
the ducting is free from leaks in case the internal pressure is below that of the
external atmosphere (see Annex C). Additional protective measures, for example,
combustible gas detectors, should be considered to ensure that the ducting is free
of flammable concentrations of gas vapor.
Additional Purge Time to Account for Ducting
The purge duration should be increased by the time necessary to purge the free
volume of those associated ducts which are not part of the equipment by at least
five times their volume at the minimum flow rate specified by the manufacturer.
Power for Protective Gas Supply
The electrical power for the protective gas supply (blower, compressor, etc.)
should be either taken from a separate power source or taken from the supply
side of the electrical isolator for the pressurized enclosure.
Enclosure Maximum Overpressure
The user should limit the pressure as specified by the manufacturer.
3-34 | Thermox WDG-V / VC / VCM
AMEVISION DISPLAY USER INTERFACE OR
COMMUNICATIONS
Purpose
The AMEVision Display User Interface provides a remote, color, graphical user
interface as well as a communications suite for the WDG-V. The AMEVision
Display User Interface provides the ability to easily configure, calibrate and
monitor up to four WDG-V sensors from a single unit. The interface between the
AMEVision Display User Interface and the WDG-V sensor is two-wire MODBUS RTU.
The AMEVision Display User Interface provides the following communication
interfaces:
•
MODBUS RTU
•
Web Interface (TCP/IP)
•
USB (Flash Drive)
Additional features:
•
4.2” Color, 1/4 VGA Display
•
19 Key Keypad
•
Intuitive Menus
•
Alarm and Event logging
•
Data logging and trending. Data can be graphed on the local display and/or
uploaded via USB flash drive
•
Upgradable firmware for attached sensors and/or AMEVision Display User
Interface USB flash drive
•
Calibration history display and plot
•
Sensor Diagnotics
AMEVision Display User Interface or Communications | 4-1
Color 1/4 VGA
Navigation
Buttons
Enter/
Menu Key
Number Pad
Help
Cancel/Escape
Figure 4-1. AMEVision Display User Interface
Reference the AMEVision Display User Interface manual for more information.
4-2 | Thermox WDG-V / VC / VCM
Setting the Analyzer Address
The interface bewteen the AMEVision Display User Interface and the WDG-V
sensor is a two-wire MODBUS RTU. Up to four WDG-V sensors can be connected to a single AMEVision Display User Interface unit. Each attached sensor
must have a unique address. The address on the WDG-V sensor is set via the dip
switch on the front of the electronics PCB.
Figure 4-2. Dip Switch Locations
321 Switch
Address
001
1
010
2
011
3
100
4
1 = ON (UP)
0 = OFF (DOWN)
AMEVision Display User Interface or Communications | 4-3
Connecting a Single WDG-V Sensor to the AMEVision
Figure 4-3. Single WDG-V Sensor to the AMEVision Display User Interface
When only one sensor is connected, the WDG-V address should be
set to “1”.
NOTE
Connecting Multiple WDG-V Sensors to the
AMEVision
Up to four sensors can be connected to a single AMEVision Display User Interface unit. The WDG-V sensor provides extra terminal connections to facilitate
easy daisy chain connections.
Figure 4-4. Multiple WDG-V Sensors to the AMEVision Display User Interface
NOTE
4-4 | Thermox WDG-V / VC / VCM
Each analyzer must have a unique address (reference address setting
section).
AMEVision Display User Interface or Communications | 4-5
When connecting multiple sensors in a daisy chain configuration, the communication termination resistor should be enable on the last sensor in the chain
(farthest from the AMEVision Display User Interface). The termination enable is
controlled by switch SW-5 of the dip switch on the front of the electronics PCB.
SW-5 on: Termination Enabled
SW-5 off: Termination Disabled
The maximum distance between the AMEVision Display User Interface and the WDG-V analyzer is 4000 feet (1219.2 meters).
NOTE
Troubleshooting
The interface between the WDG-V and the AMEVision Display User Interface is
simple and robust. However, if you experience connectivity issues, here are some
of the possible causes:
No Communications:
•
Polarity reversed
•
Address not set correctly
•
Maximum distance exceeded
•
Broken connection
Intermittent Communications:
•
Poor wiring (i.e. lack of shielded twisted pair)
•
Termination enabled on more than one unit (on multi-unit systems)
•
Termination not enabled (on multi-unit systems)
•
Maximum distance exceeded
4-6 | Thermox WDG-V / VC / VCM
WDG-V PC Configurator Software
Overview
The WDG-V PC Configurator Software provides a convient way to configure,
calibrate and monitor a single WDG-V analyzer. The interface between the configurator software and the WDG-V sensor is two-wire MODBUS RTU.
Connecting to the WDG-V
Using the optional USB to RS485 converter (P/N 1000-724-VE), attach as
shown.
The address must be set per the Settings >> Address section.
An AMEVision Display User Interface and PC software cannot be
simultaneously connected to the analyzer.
NOTE
1000-724-JE
1000-724-JE
CONNECTOR
WDG-V
TERMINAL
T/R+
T/R-
MB+
MB-
USB TO COMPUTER
Note: Only one connection can be made to the sensor.
The AmeVision and Configurator cannot be connected simultaneouslt.
AMEVision Display User Interface or Communications | 4-7
This page intentionally left blank.
4-8 | Thermox WDG-V / VC / VCM
FLOW SETTING
Setting the Flow
The recommended initial aspirator pressure setting is 3 psi. This will vary depending on the pressure/vacuum of the process for which the WDG-V is installed. If you purchased an AMEVision Display User Interface, you can use the
Flow Sensor output diagnostic screen to set the flow to the ideal rate. If you do
not have an AMEVision Display User Interface, it is recommended to adjust the
aspirator until you get a response from the analyzer, then increase the pressure by
0.25 to 0.5 psi.
Setting the flow using the AMEVision Display User Interface
Flow Sensor Diagnostic
The internal flow sensor output is displayed is percent to ideal sample loop flow.
For example, a reading of 100% indicates ideal flow whereas a reading of 50%
indicates half the ideal flow rate. The update interval for the flow sensor is 90
seconds, so you must wait at least 90 seconds for the flow output value to change.
It is recommended to operate the sensor between 80% and 120%. A low flow
alarm is generated when the flow is 45% or lower.
To set the flow, first adjust the aspirator pressure regulator until you get a response from the analyzer. Wait 90 seconds and check the flow sensor output.
Adjust the aspirator as necessary until the flow sensor reads approximately 100%.
Small increments of 0.25 psi are recommended.
Flow Setting | 5-1
This page intentionally left blank.
5-2 | Thermox WDG-V / VC / VCM
LED STATUS INDICATOR
Main Status LED (Light Pipe Visible From the Front)
OFF – Power is off.
NOTE
The status LED is NOT directly connected to the WDG-V power
supply. In the unlikely event of an error condition, it is possible for
the analyzer to be energized with dangerous voltages even though
the status LED is off. Always verify power has been removed before
servicing the instrument.
GREEN – Analyzer is in normal operating mode.
YELLOW – Analyzer is in warm-up state or diagnostic state.
RED – Analyzer has a service alarm and is in need of attention.
Figure 6-1. Analyzer is in normal operating mode.
LED Status Indicator | 6-1
Figure 6-2. Analyzer is in warm-up state or diagnostic state.
Figure 6-3. Analyzer has a service alarm and is in need of attention.
6-2 | Thermox WDG-V / VC / VCM
Printed Circuit Board (PCB) Status LED’s
There are two small, surface-mount green LED’s on the PCB located on the back
left corner of the PCB (when looking at the analyzer). You must look up into
the electronics enclosure to view these LED’s, they are not visible when looking
directly at the analyzer.
REAR LED – Heart Beat LED. Blinks once a second indicating the analyzer
software is operating. If off or not blinking this indicates the power is off or the
electronics are not functional.
FORWARD LED – Modbus Communications LED. Toggles upon receipt of a
valid Modbus message. This LED will blink rapidly in short bursts when communicating properly to the AMEVision Display User Interface.
LED Status Indicator | 6-3
This page intentionally left blank.
6-4 | Thermox WDG-V / VC / VCM
TROUBLESHOOTING
The operations in this chapter should be performed only by qualified
service personnel experienced in electrical safety techniques.
Never service the sensor unless power has been removed from the sensor, and
the sensor has been allowed to cool for at least one hour. Also, always use gloves
when working on the sensor.
This chapter describes AMEVision Display User Interface and error messages. It
also provides troubleshooting assistance. System and error messages scroll on the
bottom line of the display at three second intervals until the condition is corrected
or has ended. These system and error messages are listed in this chapter alphabetically to make them easier to locate.
NOTE
System or error messages will not appear on the display when you
are navigating the menus. However a system alarm ICON (SHOW
ICON) will appear in the top left hand corner of the screen (show
display and icon).
Troubleshooting | 7-1
General Troubleshooting
Your system may pass calibrations, yet still seem to be reading incorrect oxygen
levels. If this is the case, you may want to check the following:
Leak Check
Leaks can lead to inaccurate readings, especially if operating under a significant
pressure or vacuum.
•
Check that all compression fitting and pipe thread connections are leak tight.
•
Be sure that the mounting plate or mounting flange gasket on the rear of the
sensor is in place (see the “Sensor Mounting” section in Chapter 2 for details
on the placement of the gasket).
Sniffing for Leaks
For processes under vacuum you can check for leaks by sniffing the fittings with
another gas (for example, nitrogen or pure oxygen), being sure to avoid the area
over the top of the cell.
1. Use a piece of tygon or plastic tubing with a 1/8” stainless steel nozzle to apply the gas from a cylinder (using stainless steel prevents any problems that
might occur with plastic melting the nozzle on hot sensor components).
2. Monitor the response from the cell. When the cell millivolt reading changes,
it indicates a leak in that area of the plumbing (the vacuum of the process
pulls in the gas).
3. If not convenient to view the AMEVision display, you can also apply a voltmeter to the Cell terminals on the sensor board labeled (+O2-) to see if the
cell millivolts change, indicating a leak.
Pressurizing for Leaks
1. Remove the sensor from the process (after allowing sensor to cool) and pressurize it with 5 lbs. of air, plugging any exiting ports, the inlet probe, and
exhaust tube holes. Then go over the sensor fittings with a leak detector fluid.
2. If you see bubbles, it indicates a leak. If using this method be sure to prevent
the liquid from reaching the furnace. If the furnace does get wet, allow it sufficient time to thoroughly dry.
7-2 | Thermox WDG-V / VC / VCM
Plugged Plumbing Check
Exercise care when working on the sensor. Allow the unit to cool and
wear gloves.
1. Examine the inlet and exhaust for plugging problems. When possible, rod out
the probe and exhaust.
2. If this doesn’t solve the problem, disassemble the analyzer to locate the plug.
Clean the plumbing using hot water and a bottle brush.
3. When assembly is complete, check for leaks using a leak detecting liquid.
Aspirator Air Not Pulling Sample from Process
The aspirator will need cleaned with the Aspirator Clean Out Tool (P/N
7000-404-KE) per instructions.
Troubleshooting | 7-3
Diagnostics Checks
This section shows you how to check different sensor areas for possible problems. See the “Error Messages” section for help on what checks you should
perform based on the error message displayed. If you don’t see an error message,
yet feel your readings are inaccurate, you may also want to check the “General
Troubleshooting” section that follows. We recommend monitoring the thermocouple millivolts and cell millivolts be displayed during troubleshooting as a
troubleshooting aid. This information will be helpful should you need to contact
the factory for assistance. Be sure to always include your analyzer model and serial number when calling the factory for technical support.
Diagnostic checks are broken down as follows:
Wiring Checks
Thermocouple Checks
Calibration/Aspirator Setup Checks
AC Power Checks
Furnace Checks
Process Pressure Checks
Cell Checks
Exercise care when working on the sensor. Turn off power, allow the
unit to cool, and wear gloves.
7-4 | Thermox WDG-V / VC / VCM
Thermocouple Checks
Open Thermocouple
Shorted/Failed Thermocouple
Reversed Thermocouple Wires
Remove power to the sensor. Measure across terminals
“+TC-” on the sensor board with an ohm meter. If an open
is measured, replace the thermocouple.
Check that the thermocouple leads are not shorted to
chassis ground by using an Ohm meter to measure
between terminal TC+ on the sensor board and chassis
ground, and between terminal TC- on the sensor board
and chassis ground. If shorted, replace the thermocouple.
To verify the operation of the thermocouple itself, remove
the thermocouple from the sensor and heat its ceramic tip
to a known temperature. Measure the millivolt output with
a proper temperature indicating meter (type K thermocouple). If the thermocouple reads inaccurately, replace it.
First, view the cell temperature via the Home screen or
Diagnostic screen. If the thermocouple leads are reversed,
the displayed temperature will be decreasing as the sensor
warms up (this will usually happen at start-up or after you
replace a thermocouple). This indicates that the thermocouple wiring is reversed. If you just replaced a thermocouple, check the leads from the thermocouple to the
sensor board: yellow wire connects to terminal TC (+), red
wire connects to terminal TC (-).
Troubleshooting | 7-5
Calibration/Aspirator Setup Checks
Calibration Gas Check
To check calibration gas
values:
Check that the correct calibration gas values have been entered into the analyzer.
Select the Cal Gas Value
menu option from the Calibrate key on the control unit
and check that the calibration gas values entered
match the analyzed concentration of the cylinders.
Check that the calibration gas cylinders are turned on and are
not empty.
Calibration Line Check
Ensure that your calibration line is not contaminated with such
things as pipe dope, cutting fluid, oil, or solvents. All these
contaminants produce hydrocarbon vapors that interfere with
the proper calibration of your analyzer, resulting in lower than
expected oxygen readings. To test for contaminated lines, you
must temporarily bypass your current calibration line with a clean
calibration line (directly from cylinder to the sensor calibration
inlet port, using a flow meter to set the proper flow) and compare
the response with that from the possibly contaminated line.
The best calibration gas to
use for this check is an O2
zero calibration gas.
Remote Calibration Unit
(RCU) Problem
Check for the proper flow rate and proper delivery pressure of
calibration gases when the remote calibration unit has been
activated (see the Inject Cal Gas Diagnostic option in the
AMEVision Display User Interface manual for help).
The RCU contains one normally open solenoid-the aspirator air
solenoid. All other solenoids are normally closed. Problems with
the RCU are usually as follows:
Plumbing Leak
To check for plumbing leaks, disconnect power from the RCU
and pressurize the inlets. Apply a leak detecting liquid along
the base of the solenoids and any plumbing fittings. Repair any
leaks found.
Solenoid Not
Energizing
The solenoid drive signal is a 12 VDC signal. This is used to
close the aspirator solenoid and open the appropriate calibration gas solenoid. To test a solenoid valve, select Inject Cal Gas
from the Calibrate key menu to energize that solenoid. Verify
that the solenoid drive signal is present. The aspirator air solenoid is closed when the drive signal is applied. Other solenoids
are open when the drive signal is applied (cal gas flowing). With
the drive signal present, verify the proper flow. If no signal is
present, check the interconnecting wiring between the control
unit and the RCU.
If wiring is correct, replace the electronics. If the solenoid drive
signal is present, it indicates a problem with one of the solenoid
valves. Replace the solenoid valve, being sure that the solenoid
O-ring seals are properly positioned.
7-6 | Thermox WDG-V / VC / VCM
To check if a
calibration gas
solenoid is stuck
open:
Verify that no drive signal is present and check for flow on the
RCU flow meter. If flow is indicated when no solenoids are energized, a solenoid is stuck open. Shut off your calibration gases
(one at a time) until the flow drops to zero. This identifies the
defective solenoid valve (replace the solenoid valve, being sure
that the solenoid O-ring seals are properly positioned).
Calibration Gas If you are having problems running an auto calibration, you
Time Inadequate may not have allowed the calibration gases enough time to flow
through the sensor and stabilize. To correct this problem, select
Inject Cal Gas from the Diagnostic menu.
Turn on each calibration gas and determine how long it takes
for each gas to stabilize on the control unit display. Then add a
one minute buffer to each of these times. Also make sure the
cylinder regulators are set to the correct pressure. See the Flow
Section of the System Interconnect Drawing for the system flow
and pressure requirements. This drawing is included with your
Installation package.
Select Cal Gas Duration from the Calibration menu to set calibration gas times.
NOTE
If performing a manual calibration, ensure that you are waiting for the
reading to stabilize on the display before switching to the next calibration gas.
AC Power Checks
Loss/Inadequate AC Voltage Measure the AC voltage to the sensor board at terminals L and
to the Sensor
N. Ensure that this voltage is sufficient. Check the measurement technique used by the Volt meter (for example, RMS,
average, peak, etc.). Specifications are based on RMS measurements.
Furnace Checks
Open Furnace
Disconnect power to the control unit and the sensor. With Ohm
meter, measure across Terminals “FURN” on the sensor board.
The resistance of the furnace should be:
32 ohms (±10%)
If the furnace resistance is not within allowable tolerances,
replace the furnace.
Loss of AC Power
to the Furnace
Verify the correct line voltage at L and N of the sensor board.
With line voltage present, check the voltage at terminals FURN
on the sensor board. If voltage is present and the system is not
heating, remove power and check the furnace resistance (see
“Furnace Checks” section). If no voltage is present at terminals
FURN on the sensor board, replace the electronics.
Troubleshooting | 7-7
Process Pressure Checks (Only if above 2 psig)
To check that you entered the process pressure correctly, select Process Pressure from the
configuration menu. Also ensure that the calibration process pressure equals normal operating process pressure. Calibration should only be performed under these conditions for highest
accuracy.
Cell Checks
If the cell fails when you first begin
to use the analyzer:
It is likely that there is a leak in the sensor plumbing
or an improper calibration gas setup, and there is not
a problem with the cell itself (see the “Leak Check”
section later in this chapter for help on how to check
for leaks; see the “Calibration/Aspirator Air Checks”
section earlier in this chapter for help on checking
your calibration gas setup).
If the analyzer has been operating
for some time and you feel the oxygen reading is inaccurate:
First check by running a known calibration gas to verify the analyzer’s response. If the analyzer responds to
the calibration gas correctly, it indicates either leaking
or plugged plumbing.
If your analyzer doesn’t respond
properly to the calibration gas:
This may indicate a problem with the cell. Before replacing the cell, check for leaks or plugged plumbing.
7-8 | Thermox WDG-V / VC / VCM
Alarm and Warning Messages
ALARMS
ALARM
DESCRIPTION
Possible Causes
Cell T/C Failure
Thermocouple is not connected or is open
Bad T/C, Bad connection
Cold Junction Compensator
Failure
The PCB temperature sensor has failed
Bad Electronics
Cell Temperature Control
A critical overtemperature condition has
occurred
Shorted, open, or Bad
T/C or Failed Temperature
Control Circuit.
Unit must be reset to clear
this alarm
Cell Over Temp
Cell temperature is 4 degrees or higher
above the setpoint
Bad Thermocouple, Failed
temperature control circuit
Cell Under Temp
Cell temperature is 4 degrees or lower
below the setpoint
Faulty heater, faulty heater
connection, Faulty T/C,
Faulty T/C connection,
Low Power.
Only after warm-up is
complete
Cell Temp Rise Failure
Cell heater failed to heat the cell
Bad Cell furnace heater,
Bad heater connection,
Shorted T/C, Bad T/C connection.
Unit must be reset to clear
this alarm
Over Temp Relay Tripped
The cell over-temperature relay has tripped Open Thermocouple, Over
Temperature condition,
Failed Temperature control
circuit.
Must cycle power to reset
the relay
RTD Failure
The Box Temp RTD Failed
Open or Shorted RTD
Box Temp High
Box temperature is 4 degrees or higher
above the setpoint
Failed temperature control
circuit
Box Temp Low
Box temperature is 4 degrees or lower
below the setpoint
Faulty heater, faulty heater
connection, Faulty RTD,
Faulty RTD connection,
Low Power.
Only after warm-up time
expires
Box Temp Rise Failure
Box temperature failed to heat
Bad box heater(s), Bad
heater connection, Bad
RTD, Bad RTD connection
Troubleshooting | 7-9
Low Sample Flow
Low sample flow detected
Loss of aspirator air
plugged inlet probe,
plugged sample system.
Active only after warm-up
time expires
Cell Failure (Open)
Cell resistance exceeds the normal limit
Bad Cell, Open Cell Connection
Combustible Detector Open
Comb detector is open or not connected
Bad Detector, Bad electronics.
Only when configured as
WDG-VC
Methane Detector Open
CH4 detector is open or not connected
Bad Detector, Bad electronics.
Only when configured as
WDG-VCM
Cell Mv Mismatch
The cell mv measurement does not match
the real time cell meas
Bad A/D circuit (Bad electronics)
Analog Output 1 Error
The analog output measurement does not
match the setting
Open 4-20mA loop or
failed output or bad readback circuit
Flow Sensor Failure
The flow sensor failed (open)
Bad flow sensor or circuit
Cell T/C Measurement Mismatch
The Redundant temperature measurement
does not match the real time meas
Bad T/C or circuit
7-10 | Thermox WDG-V / VC / VCM
WARNINGS
Cell Life Nearing Its End
The ZrO2 cell is nearing its end of life. This
is based on calibration history.
Comb Detector Life Nearing
Its End
The combustible detector sensitivity is near Low detector output
its end of life. This is based on calibration
history.
Methane Detector Life Nearing The methane detector sensitivity is near
Its End
its end of life. This is based on calibration
history.
High Cell Mv
Low cell output
Low Detector output
The combustible measurement is no longer Hydrocarbon upset
valid (set to full scale) due to lack of O2
Analog Output 1 Out Of Range The analog output value is saturated (high
or low)
Improperly set range
Analog Ouptut 2 Out Of Range The analog output value is saturated (high
or low)
Improperly set range
Analog Ouptut 3 Out Of Range The analog output value is saturated (high
or low)
Improperly set range
Last O2 Span Calibration
Failed
The last oxygen span calibration failed
Tank Empty, bad solenoid,
didn’t wait long enough,
Bad Cell
Last O2 Zero Calibration Failed The last oxygen zero calibration failed
Tank Empty, bad solenoid,
didn’t wait long enough,
Bad Cell
Last Comb Calibration Failed
The last combustible calibration failed
Tank Empty, bad solenoid,
didn’t wait long enough,
bad detector
Last CH4 Calibration Failed
The last methane calibration failed
Tank Empty, bad solenoid,
didn’t wait long enough,
bad detector
Last Flow Calibration Failed
The last flow sensor calibration failed
Plugged sample system or
bad sensor
Oxygen Calibration Required
O2 Calibration is required
Last Calibration failed or
unit has never been calibrated
Comb Calibration Required
Combustible Calibration is required
Last Calibration failed or
unit has never been calibrated
Methane Calibration Required
CH4 Calibration is required
Last Calibration failed or
unit has never been calibrated
Troubleshooting | 7-11
This page intentionally left blank.
7-12 | Thermox WDG-V / VC / VCM
SERVICE AND PARTS
The operations in this appendix should be performed only by qualified service personnel with a knowledge of electrical safety techniques.
Never service the sensor unless power has been removed from the
sensor, and the sensor has been allowed to cool for one hour.
This chapter is divided into the following sections:
•
Cell Replacement
•
Thermocouple Replacement
•
Furnace Replacement
•
Combustibles/Methane Detector Replacement (optional)
•
Flow Sensor Replacement
•
Box Heater Wiring
•
RCU (Remote Calibration Unit) Solenoid Valve Replacement
•
Electronics Board
•
Recommended Maintenance Schedule
•
Replacement Parts List
You can clean the outside of the sensor using normal household or commercial
general purpose cleaners, cloths, or sponges. You can also use water. Always turn
off power before attempting to clean the enclosure.
A wiring diagram for a standard sensor is provided in Figure 8-1. This figure
shows the locations of all wiring connections on the sensor. If your sensor requires different wiring, this is provided in an appendix.
Service and Parts | 8-1
Warnings
•
If you turn off the analyzer or the process is shut down, turn off the aspirator
air to avoid plugging problems.
•
The outside of the sensor cover and all sensor assembly components are
extremely hot (up to 500°F (260°C) inside the cover), even after a considerable period from shutdown. Turn off the power to the sensor and control unit
when working inside the sensor. Use caution and wear appropriate gloves
when handling components.
•
Be extremely careful when performing maintenance on the sensor while the
process is running, especially if the process is under significant positive pressure. Removing any part of the sensor plumbing can allow process gases and
gases of high temperature to escape into the sensor.
•
Always use a backup wrench when working on sensor plumbing. This helps
to prevent damaging welds and distorting sensor plumbing.
•
Never use pipe dope or any other contaminant that gives off combustible
vapor on any joints of the sample tubing. Combustible vapor in the sample
tubing can lead to erroneous readings.
Cautions
8-2 | Thermox WDG-V / VC / VCM
METHANE DETECTOR
REFERENCE ELEMENT
RTD
METHANE DETECTOR
ACTIVE ELEMENT
CELL CELL +
THERMOCOUPLE +
THERMOCOUPLE SIL THERMOCOUPLE +
SIL THERMOCOUPLE COMBUSTIBLES DETECTOR
ACTIVE ELEMENT
COMBUSTIBLES DETECTOR
REFERENCE ELEMENT
350W HEATER AC
LINE
NEUTRAL
CUSTOMER CONNECTIONS
50W HEATER AC
(HALF WAVE RECTIFIED)
CELL FURNACE AC
FLOW SENSOR SENSE ELEMENT
FLOW SENSOR HEATER ELEMENT
Figure 8-1a. Wiring diagram for standard sensor.
Service and Parts | 8-3
SPARE
PROCESS ALARM
(ALARM 5)
PROCESS ALARM
(ALARM 4)
ASPIRATOR
SOLENOID
ZERO GAS
VALVE
SPAN 1
CONTROL
SPAN 2
ANALOG OUTPUT
ANALOG INPUT
SENSOR MODBUS
MODBUS (DAISY CHAIN)
LINE
NEUTRAL
GND SCREW
CONTACT INPUT
FUTURE
ANALOG OUTPUT
ANALOG OUTPUT
24 TO 36VDC ANALOG OUTPUT
EXTERNAL VOLTAGE SUPPLY
AC PWR
SUPPLY
8-4 | Thermox WDG-V / VC / VCM
SERVICE ALARM
(ALARM 1)
DATA VALID ALARM
(ALARM 2)
PROCESS ALARM
(ALARM 3)
SPARE
12VDC FOR RCU SOLENOID DRIVE
Figure 8-1b. Customer wiring diagram for standard sensor.
Cell Replacement/Cleaning
Disconnect power from the sensor and control unit and allow the
sensor to cool before replacing parts inside the sensor.
Use Figure 8-2:
1. Open the sensor cover to expose sensor components.
2. Remove the cell clips.
3. While holding the top of the cell housing with a backup wrench, loosen the
lower hex nut (not the top hex nut), then remove the entire cell assembly.
(See the * in Figure 8-2).
Loosening the cell may require a good deal of torque.
NOTE
4. If cleaning the cell, wash with water or alcohol. Dry the cell thoroughly before reinstalling. Always use a new cell O-ring.
If replacing the cell, discard the old cell and cell O-ring and retrieve the new
cell with supplied cell O-ring. Avoid touching the bare cell. Instead, hold the
cell by one of its hex nuts.
5. Place the cell O-ring on the base of the cell as shown in Figure 8-2.
6. Insert the new cell into the cell housing. Do not touch the bare cell when
reinserting. Tighten the cell into the cell housing using the lower hex nut
(the upper hex nut is pre-tightened at the factory). The cell O-ring provides a
seal for the system. Make sure it is aligned into its proper recess and evenly
crushed when tightening the cell.
Service and Parts | 8-5
FURNACE
CELL HOUSING
O-RING
CELL
HEX NUT
WASHER
*
HOLD CELL HOUSING
USING 7/8" OR 1" WRENCH
(23mm or 26mm)
TURN HEX NUT COUNTER
CLOCKWISE USING A
7/8" WRENCH (23mm)
CELL CLIPS
Figure 8-2. Cell replacement.
CELL REPLACEMENT
8-6 | Thermox WDG-V / VC / VCM
Thermocouple Replacement
The tip of the thermocouple is positioned in the furnace so that it is
near, but not touching, the cell housing or the furnace heater coil.
NOTE
Use Figure 8-3:
1. Open the sensor door to expose sensor components.
2. Disconnect thermocouple wires from terminals +TC- on the sensor board.
3. Remove the 2 speed clips from the thermocouple mounting tabs. Use needlenose pliers to remove the clips.
4. Pull the thermocouple straight down and remove it from the furnace.
5. Insert the new thermocouple aligning the tab with the two posts on the bottom plate. Be careful not to damage the ceramic tip while inserting the assembly.
6. Attach one new speed nut to each post.
7. Attach the thermocouple wires to the +TC- terminals on the sensor card (yellow wire is positive (+)).
8. Perform an oxygen calibration.
Service and Parts | 8-7
Figure 8-3. Thermocouple replacement
8-8 | Thermox WDG-V / VC / VCM
Combustibles/Methane Detector Replacement
The Combustible detector will always face outward, and the Methane
detector will always face inward from the factory.
NOTE
Combustibles Detector Replacement
Use Figure 8-4:
1. Disconnect the combustible detector (Black and Brown) wires on the sensor
board.
2. Loosen the two 4.0 mm hex screws that holds the combustible detector in place
and remove the detector from its housing. Save the screws from the detector to
use when installing the new combustibles detector.
3. Always use a new O-ring when detector is removed.
4. Install the new combustible detector in place, aligning the two screws on the
block. Be careful not to damage the detector element ends while inserting the
assembly. Tighten screws until face is flush with the block and gasket is fully
compressed.
5. Reconnect the detector leads to the sensor board.
Methane Detector Replacement
1. You must remove the block in order to access and replace the Methane detector.
2. Disconnect the methane detector (Blue and White) wires on the sensor board.
3. Loosen the two 4.0 mm hex screws that holds the methane detector in place
and remove the detector from its housing. Save the screws from the detector to
use when installing the new combustibles detector.
4. Always use a new O-ring when detector is removed.
5. Install the new methane detector in place, aligning the two screws on the
block. Be careful not to damage the detector element ends while inserting the
assembly. Tighten screws until face is flush with the block and gasket is fully
compressed.
6. Reinstall the block.
7. Reconnect the detector leads to the sensor board.
Service and Parts | 8-9
Figure 8-4. Combustibles detector replacement
8-10 | Thermox WDG-V / VC / VCM
Flow Sensor Replacement
Flow Sensor Replacement
Use Figure 8-5:
1. Disconnect the flow sensor Red wires (Heater) and the Yellow wires (Sense)
on the sensor board.
2. Loosen the nut that connects the heater block and the flow sensor.
3. Loosen the nut that connects the furnace and the flow sensor. Brace the furnace assembly by using a second wrench attached to the sensor nut under the
furnace. See Figure 8-5 for reference.
4. Remove the flow sensor.
Handle the new flow sensor with care. It is very delicate when not
installed in the system.
NOTE
5. Install the new flow sensor in place. Tighten the two connecting nuts.
6. Reconnect the Red wires (Heater) and the Yellow wires (Sense) on the sensor
board.
Service and Parts | 8-11
Figure 8-5. Flow sensor replacement
8-12 | Thermox WDG-V / VC / VCM
Furnace Replacement
Furnace Replacement
Use Figure 8-6:
1. Disconnect cell furnace wires.
2. Remove the Thermocouple assembly from the furnace.
3. Remove the Cell.
4. Loosen the swage nut at the top and bottom of the furnace. You may need to
remove the flow sensor to get the furnace assembly out.
5. Remove the Furnace assembly.
6. Install the new Furnace assembly in place.
7. Reconnect the Thermocouple, Flow Sensor and Cell.
8. Reconnect the wires on the sensor board.
Figure 8-6. Furnace Assembly
Service and Parts | 8-13
Heater Replacement
There is only one heater in the O2 version. There are two heaters in
the Combustible/Methane version of the WDG-V.
NOTE
Box Heater Replacement
Use Figure 8-7:
1. Disconnect heater wires.
2. Remove the speed clip from the mounting tab. Use needle-nose pliers to
remove the clip.
3. Remove the box heater by sliding it out.
4. Replace the box heater
5. Attach a new speed nut to the post.
6. Reconnect the wires on the sensor board.
8-14 | Thermox WDG-V / VC / VCM
Figure 8-7. Box Heater Replacement
Service and Parts | 8-15
Combustible Block Heater Replacement
(Only on Combustible/Methane Units))
Use Figure 8-8:
1. Disconnect heater wires.
2. Loosen the 4.0 mm hex screw.
3. Remove the combustible block heater by sliding it out.
4. Replace the box heater
5. Tighten the hex screw into place.
6. Reconnect the wires on the sensor board.
8-16 | Thermox WDG-V / VC / VCM
Figure 8-8. Combustible Block Heater Replacement
Service and Parts | 8-17
Electronics Replacement
Electronics Replacement
Use Figure 8-9:
1. Disconnect all wires.
2. Loosen the two Phillips screw located on the back of the electronics plate.
3. Slide the board out until the pegs at the top line up with the holes in the electronics board plate and drop it down.
2. Replace the electronics board.
2. Properly tighten the screws.
8-18 | Thermox WDG-V / VC / VCM
Figure 8-9. Electronics Board Removal
Service and Parts | 8-19
Recommended Maintenance Schedule
Calibration
Check calibration and/or recalibrate the analyzer every 90 days.
8-20 | Thermox WDG-V / VC / VCM
Replacement Parts List
Sensor
Assy, Furnace, WDG-V, 120 VAC
Assy, Furnace, WDG-V, 240 VAC
Kit, WDG-V Replacement Thermocouple For O2 Cell Furnace
Combustibles Detector Kit - RTD Detector
Combustibles Detector Kit - Hot Wire Detector
Methane Detector Kit
WDG-V Sensor Main Processor Board For RTD Comb. Detector
WDG-V Sensor Main Processor Board For Hot Wire Comb. Detector
Assy, WDG-V Boards and Plate For RTD Comb. Detector
Assy, WDG-V Boards and Plate For Hot Wire Comb. Detector
Kit, Combustibles Block w/Nut and Ferrules
Kit, Combustibles/Methane Block w/Nut and Ferrules
Kit, Flow Sensor, WDG-V
Cell O-Ring Metal
Kalrez O-Ring, Combustibles/Methane Detector Housing
Standard Zirconia Cell
Severe Service Zirconia Cell - Consult Factory For Application
Kit, WDG-V Aspirator w/ Nut And Ferrules
Flashback Assy, 1” Element
Kit, Convection Loop w/ Nut and Ferrules, O2 Only
(long tube for use without flame arresters)
Kit, Flame Arrester Convection Loop w/ Nut and Ferrules
(short tube for use with flame arresters)
Cell Clip Assembly - Inside
Cell Clip Assembly - Outside
Assy, WDG-V, Box Heater, 120VAC
Assy, WDG-V, Box Heater, 240VAC
Heater Plate WDG-V
Kit, Cartridge Heater 350W, 120V, Includes Heater and Push Nut
Kit, Cartridge Heater 50W, 120V, Includes Heater and Screw
Kit, Cartridge Heater 350W, 240V, Includes Heater and Push Nut
Kit, Cartridge Heater 50W, 240V, Includes Heater and Screw
Kit, WDG-V Replacement 115 VAC Heater Coil and Push Nut
Kit, WDG-V Replacement 240 VAC Heater Coil and Push Nut
RTD - Box Temp Control - 100 Ohm (O2 Only)
RTD - Box Temp Control - 100 Ohm (Comb/Methane)
Threaded Exhaust Tube
Sample Probe
7000-719-SE
7000-720-SE
7000-839-TE
7000-815-TE *
7000-816-TE **
7000-814-TE
8000-119-SE
8000-148-SE
7000-845-SE ***
7001-214-SE ***
7001-226-TE****
7001-227-TE****
7000-817-TE
42005JE
3000-337-JE
7000-568-SE
7000-733-SE
7001-216-TE
71212SE
7001-224-TE
7001-225-TE
72329SE
72328SE
7000-843-SE *****
7000-844-SE *****
7000-405-KE
7000-846-TE
7000-848-TE
7000-847-TE
7000-849-TE
7000-840-TE
7000-841-TE
25384JE
1000-630-JE
70619KE
by serial number
Service and Parts | 8-21
Remote Calibration Unit (RCU)
Air Regulator
Regulator Gauge
Aspirator Air Solenoid Valve (normally open)
Calibration Gas Solenoid Valve (normally closed)
Flow Meter
39004JE
37070JE
36090JE
36088JE
37020JE
* The RTD detector is the standard detector for most appications.
Requires sensor main processor board 8000-119-SE
** The Hotwire detector is the severe service detector. Consult factory for
applications. Requires sensor main processor board 8000-148-SE
*** Sensor assembly includes I/O connection board, main processor
board, and mounting plate
**** The block heater, RTD, and detector are not included
***** Includes heat plate, heater, push nut, and dowel pin
When ordering, provide the serial number of your analyzer to ensure proper parts are
ordered:
AMETEK
Process & Analytical Instruments Division
150 Freeport Road
Pittsburgh, PA, USA 15238
Phone: (412) 828-9040 Fax: (412) 826-0399
8-22 | Thermox WDG-V / VC / VCM
DRAWINGS AND CUSTOM INSTRUCTIONS
This appendix provides any custom drawings or instructions you may have
ordered in addition to the standard WDG-V analyzer. If you didn’t order any
custom options, the standard Interconnect Drawing is provided.
If you did order any special options, the drawings or special instructions provided
here supercede any drawing or options provided elsewhere in this manual.
Drawings and Custom Instructions
| 9-1
This page intentionally left blank.
9-2 | Thermox WDG-V / VC / VCM
CUSTOMER MODBUS MAP
Sensor
Host
Register
Name
Data Type
Description
1
0
System Alarm
Int 32
System alarms reported by the AMEVision
Display User Interface Software
1
2
Oxygen
Float
Oxygen concentration in percentage
Writable
1
4
Combustibles
Float
Combusitble concentration in ppm
Writable
1
6
Methane
Float
Methane concetration in percentage
Read Only
1
8
Cell Temperature
Float
Cell temperature in degree C
Read Only
1
10
Box Temperature
Float
Box temperature in degree C
Read Only
1
12
Cell mV
Float
Oxygen cell in millivolts
1
14
CJ Temperature
Float
Electronics Temperature (cold junction compensator temp) in degree C
1
16
Process Alarm Mask
Unsigned Int 32 Process alarm bit mask
Writable
1
18
Command
Unsigned Int 16 Command to the analyzer
Writable
1
19
Command Response Unsigned Int 16 Response to a command
1
20
Oxygen Span Gas
Float
Oxygen span gas value
1
22
Zero Gas
Float
Zero gas value
Read Only
1
24
Combustible Span
Gas
Float
Combustible span gas value
Read Only
1
26
Methane Span Gas
Float
Methane span gas value
Read Only
1
28
Data Valid
Unsigned Int 16 Datavalid (1 if valid, 0 if not)
Read Only
1
29
State
Unsigned Int 16 Current state of the analyzer
Read Only
1
30
Event Mask 1
Unsigned Int 32 Event Mask 1
Writable
1
32
Event Mask 2
Unsigned Int 32 Event Mask 2
Read Only
1
34
Pressure
Float
Process Pressure
Read Only
Writable
Read Only
Writable
Writable
Customer Modbus Map
| 10-1
This page intentionally left blank.
10-2 | Thermox WDG-V / VC / VCM
FRENCH WARNINGS /
FRANÇAIS AVERTISSEMENTS
NOTES SUR LA SECURITE
LES AVERTISSEMENTS, LES PRÉCAUTIONS ET LES NOTES CONTENUS DANS CE MANUEL ATTIRENT L’ATTENTION SUR DES INSTRUCTIONS CRITIQUES COMME SUIT :
Une procédure de fonctionnement qui n’est pas strictement suivie
peut engendrer une blessure de personne ou une contamination de
l’environnement.
Une procédure de fonctionnement qui n’est pas strictement suivie
peut endommager l’équipement.
Informations importantes à ne pas laisser passer :
NOTE
Risque de brûlure. Surface brûlante. Ne pas toucher, laisser refroidir
avant tout travail.
French Warnings / Français Avertissements | A-1
Sécurité électrique
Une tension jusqu’à 5 kV peut être présente dans le boîtier. Coupez toujours les
alimentations électriques avant d’effectuer une maintenance ou un dépannage. Seuls
des techniciens qualifiés électriquement doivent effectuer les connexions électriques
et les vérifications de terres.
Une utilisation quelconque de l’équipement de manière non spécifiée par le constructeur peut compromettre la protection de la sécurité d’origine fournie par
l’équipement.
Mise à la terre
La mise à la terre de l’instrument est obligatoire. Les spécifications de performance
et la protection de sécurité sont invalides si le fonctionnement de l’instrument est
effectué depuis une alimentation électrique à la terre impropre.
Vérifiez la continuité de la terre de tout équipement avant
d’alimenter électriquement.
A-2 | Thermox WDG-V / VC / VCM
Note importante aux utilisateurs
Il n’y a pas de composant à maintenir à l’intérieur même des composants de
l’analyseur WDG-V. Ne pas retirer le capot d’un analyseur WDG-V. Se référer à
du personnel qualifié pour la maintenance.
Ce qui suit s’applique aux modèles WDG-V Division 2 :
L’alimentation électrique, les entrées et sorties (E/S) doivent être en accord avec
les méthodes de câblage en zone Class I Division 2 (se référer au code électrique
national) et en accord avec l’autorité locale ayant juridiction.
AVERTISSEMENT – RISQUE D’EXPLOSION – LA SUBSTITUTUON DE COMPOSANTS PEUT COMPROMETTRE LA CONFORMITE DE ZONE CLASS I DIVISION 2.
AVERTISSEMENT – RISQUE D’EXPLOSION – EN ZONE EXPLOSIBLE, COUPEZ L’ALIMENTATION ELECTRIQUE AVANT DE
REMPLACER OU CABLER UN MODULE.
AVERTISSEMENT – RISQUE D’EXPLOSION – NE DECONNECTEZ PAS L’EQUIPEMENT SAUF SI L’ALIMENTATION ELECTRIQUE A ETE COUPEE OU SI LA ZONE EST SANS RISQUE.
CET EQUIPEMENT EST CONFORME A UNE UTILISATION EN
ZONE CLASS I, DIVISION 2, GROUPES A, B, C ET D OU HORS
ZONE EXPLOSIBLE SEULEMENT.
La température ambiante maximum ce cet analyseur est de 60°C (140°F).
L’unité de terrain WDG-V et son afficheur sont des pieces complexes qui doivent
être maintenues uniquement par des techniciens qualifies ayant une expertise en
technologie de l’instrument et en systèmes électriques. AMETEK recommande
que tout analyseur nécessitant un dépannage soit renvoyé à l’usine. Vous devriez
seulement essayer de travailler sur cet équipement après avoir reçu une formation
d’un représentant AMETEK / Division P&AI. Si vous décidez de travailler sur
cet analyseur, ayez conscience que des hautes tensions, des hautes températures et
d’autres conditions potentiellement à risque peuvent surgir.
French Warnings / Français Avertissements | A-3
Sommaire de Sécurité Générale
Lisez attentivement les précautions de sécurité suivantes pour éviter de se blesser
et d’endommager l’équipement ou tout produit connecté à ce dernier.
Utilisez un câblage adéquat
Pour éviter les risques de feu, utilisez seulement le câblage spécifié dans le chapitre
Installation de ce manuel de l’utilisateur.
Evitez les surcharges électriques
Pour éviter une électrocution ou un risque de feu, n’alimentez pas électriquement
des bornes qui sont hors de la gamme spécifiée pour ces bornes.
Mettez à la terre l’analyseur
Suivez les instructions de mise à la terre fournies dans le chapitre Installation de
ce manuel de l’utilisateur. Avant de connecter aux bornes d’entrée ou de sortie de
l’analyseur, assurez-vous que l’analyseur est correctement mis à la terre.
Ne pas faire fonctionner sans capot
Pour éviter une électrocution ou un risque de feu, ne faites pas fonctionner l’analyseur
sans capot ou avec un panneau retiré.
Utilisez des fusibles adéquats
Pour éviter tout risque de feu, utilisez des fusibles de types et classes spécifiés pour
cet analyseur.
Ne faites pas fonctionner en atmosphère explosible
Pour éviter toute blessure ou risque de feu, ne faites pas fonctionner l’analyseur
dans une atmosphère explosible, sauf si vous avez acheté les options spécifiquement
conçues pour ces environnements.
A-4 | Thermox WDG-V / VC / VCM
Precautions aux Dommages de L’Analyseur
Utilisez une alimentation électrique adéquate
Ne faites pas fonctionner l’analyseur avec une alimentation électrique
qui fournit plus que la tension électrique spécifiée.
Ne faites pas fonctionner avec les défaillances suspectes
Si vous suspectez que l’analyseur est endommagé, demandez à un
technicien qualifié de l’inspecter.
Prenez des précautions en levant l’instrument
L’analyseur pèse environ 25 kg. Prenez des précautions lorsque vous le levez de
sa caisse.
Utilisez des vêtements adaptés
L’analyseur est brûlant, l’utilisateur doit porter des gants de protection lorsqu’il
manipule l’analyseur.
French Warnings / Français Avertissements | A-5
NOTE
Les opérations dans ce chapitre doivent être effectuées seulement par
un technicien qualifié et expérimenté en techniques de sécurité. Il n’y
a pas de composant à maintenir à l’intérieur même des composants
de l’analyseur WDG-V. Ne jamais travailler sur l’analyseur sans
que l’alimentation électrique ait été coupée et que l’analyseur ait pu
refroidir pendant au moins une heure. Utilisez aussi des gants si vous
travaillez sur l’analyseur.
Ne pas installer le chauffage de la sonde si le gaz échantillon
contient potentiellement un mélange explosive de combustibles
; le chauffage de la sonde peut chauffer le gaz jusqu’à son point
d’inflammation.
Ne pas ouvrir une porte à charnière de l’unité de terrain purgée Z
avant d’avoir vérifié que la zone n’est pas classée à risque.
Des pièces internes chaudes sont au-dessus de la température à
laquelle s’enflamment les gaz combustibles. L’alimentation électrique doit être coupée sur les unités de terrain et de contrôle pendant 90 minutes en maintenant le debit d’air de purge avant d’ouvrir
la porte, sauf s’il est prouvé que la zone est sure.
L’alimentation électrique ne doit pas être remise après ouverture de
l’enceinte jusqu’à ce qu’elle ait été purgée pendant 65 minutes à une
pression de 10 mm H2O.
N’ouvrez pas l’unité de calibration déportée purgée Z jusqu’à ce
que vous verifiez que la zone soit classée sans risque. Maintenez
aussi la purge Z à la pression et au debit indiqués sur la plaque
d’avertissement de l’unité.
A-6 | Thermox WDG-V / VC / VCM
Coupez l’alimentation électrique du secteur avant d’effectuer un
câblage.
Ne placez pas les fils d’alimentation électrique de l’unité de contrôle
dans les mêmes câbles que d’autres fils d’alimentation électrique. En
séparant ces fils, vous empéchez des signaux transitoires d’atteindre
l’unité de contrôle.
Dans un but de compatibilité électromagnétique, en aucune circonstance vous ne devez laisser l’écran d’un câble déconnecté en une ou
deux extrémités (unité de contrôle ou de terrain ou autre appareil).
Les operations dans ce chapitre doivent être effectuées seulement
par un technicien qualifié et expérimenté en techniques de sécurité
électrique.
Soyez prudent lorsque vous travaillez sur l’analyseur. Laissez
l’analyseur refroidir et portez des gants.
Coupez l’alimentation électrique des unités de terrain et de contrôle
et laissez l’analyseur refroidir avant de remplacer des pieces dans
l’analyseur.
French Warnings / Français Avertissements | A-7
This page intentionally left blank.
A-8 | Thermox WDG-V / VC / VCM