Yokogawa Instruction Manual Model SC450G [Style: S2] Conductivity / Resistivity Converter
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
User’s
Manual
Model SC450G [Style: S2]
Conductivity / Resistivity
Converter
IM 12D08N05-01E
4th Edition
Note
This page may be referred to when
reading pages where subsequent
submenu screens are shown in
the text. Connection to the relevant
submenu screen is indicated by a
doted line with an arrow.
Note that screens in the text
are typical examples and actual
screens may differ depending on
the set parameters.
Commissioning
Connection to the relevant submenu screen
is indicated by a doted line with an arrow.
Measurement setup
Measure Conductivity only
Configure sensor
Temperature settings
Temp. Compensation
Calibration settings
Concentration
Enter
PREFACE
Electrostatic discharge
The EXAxt converter contains devices that
can be damaged by electrostatic discharge.
When servicing this equipment, please observe
proper procedures to prevent such damage.
Replacement components should be shipped in
conductive packaging. Repair work should be
done at grounded workstations using grounded
soldering irons and wrist straps to avoid electrostatic discharge.
DANGER
Installation and wiring
The EXAxt converter should only be used with
equipment that meets the relevant IEC, American or Canadian standards. Yokogawa accepts
no responsibility for the misuse of this unit.
CAUTION
The Instrument is packed carefully with shock
absorbing materials, nevertheless, the instrument may be damaged or broken if subjected
to strong shock, such as if the instrument is
dropped. Handle with care.
WARNING
• Do not use an abrasive or organic solvent in
cleaning the instrument.
• Do not modify the SC450G converter.
• Substitution of components may impair
suitability for Division 2.
Do not remove or replace while circuit is live
unless area is know to be non-hazardous.
Explosion Hazard – Do not disconnect
equipment unless area is known to be nonhazardous.
Do not reset circuit breaker unless power
has been removed from the equipment or
the area is known to be non-hazardous.
Notice
• This manual should be passed on to the end
user.
• The contents of this manual are subject to
change without prior notice.
• The contents of this manual shall not be
reproduced or copied, in part or in whole,
without permission.
• This manual explains the functions
contained in this product, but does not
warrant that they are suitable the particular
purpose of the user.
• Every effort has been made to ensure
accuracy in the preparation of this manual.
However, when you realize mistaken
expressions or omissions, please contact the
nearest Yokogawa Electric representative or
sales office.
• This manual does not cover the special
specifications. This manual may be left
unchanged on any change of specification,
construction or parts when the change does
not affect the functions or performance of
the product.
• If the product is not used in a manner
specified in this manual, the safety of this
product may be impaired.
Yokogawa is not responsible for damage to the
instrument, poor performance of the instrument
or losses resulting from such, if the problems
are caused by:
• Improper operation by the user.
• Use of the instrument in improper
applications
• Use of the instrument in an improper
environment or improper utility program
• Repair or modification of the related
instrument by an engineer not authorized by
Yokogawa.
Safety and Modification Precautions
• Follow the safety precautions in this manual
when using the product to ensure protection
and safety of the human body, the product
and the system containing the product.
How to dispose the batteries:
This is an explanation about the new EU Battery Directive (DIRECTIVE 2006/66/EC). This
directive is only valid in the EU. Batteries are
included in this product. Batteries incorporated
into this product cannot be removed by yourself. Dispose them together with this product.
When you dispose this product in the EU,
contact your local Yokogawa Europe B.V.office.
Do not dispose them as domestic household
waste.
Battery type: silver oxide battery
Notice:
The symbol (see above) means they shall be
sorted out and collected as ordained in ANNEX
II in DIRECTIVE 2006/66/EC.
ii
The following safety symbols are used
on the product as well as in this manual.
DANGER
This symbol indicates that an operator
must follow the instructions laid out in this
manual in order to avoid the risks, for the
human body, of injury, electric shock, or
fatalities. The manual describes what special care the operator must take to avoid
such risks.
WARNING
This symbol indicates that the operator
must refer to the instructions in this manual
in order to prevent the instrument (hardware) or software from being damaged, or
a system failure from occurring.
CAUTION
This symbol gives information essential for
understanding the operations and functions.
Note!
This symbol indicates information that
complements the present topic.
This symbol indicates Protective Ground
Terminal
This symbol indicates Function Ground
Terminal (Do not use this terminal as the
protective ground terminal.)
This symbol indicates Alternating current.
This symbol indicates Direct current.
Warranty and service
Yokogawa products and parts are guaranteed free from defects in workmanship and
material under normal use and service for
a period of (typically) 12 months from the
date of shipment from the manufacturer.
Individual sales organisations can deviate
from the typical warranty period, and the
conditions of sale relating to the original
purchase order should be consulted. Damage caused by wear and tear, inadequate
maintenance, corrosion, or by the effects
of chemical processes are excluded from
this warranty coverage.
In the event of warranty claim, the defective goods should be sent (freight paid)
to the service department of the relevant
sales organisation for repair or replacement (at Yokogawa discretion). The following information must be included in the
letter accompanying the returned goods:
• Part number, model code and serial
number
• Original purchase order and date
• Length of time in service and a
description of the process
• Description of the fault, and the
circumstances of failure
• Process/environmental conditions that
may be related to the failure of the
device.
• A statement whether warranty or nonwarranty service is requested
• Complete shipping and billing
instructions for return of material, plus
the name and phone number of a
contact person who can be reached for
further information.
Returned goods that have been in contact
with process fluids must be decontaminated/disinfected before shipment. Goods
should carry a certificate to this effect, for
the health and safety of our employees.
Material safety data sheets should also be
included for all components of the processes to which the equipment has been
exposed.
iii
TABLE OF CONTENTS
PREFACE
1. INTRODUCTION AND GENERAL DESCRIPTION......................................................1
1-1. Instrument check....................................................................................................1
1-2. Application..............................................................................................................1
2. GENERAL SPECIFICATIONS OF EXAxt SC450G......................................................2
3. INSTALLATION AND WIRING......................................................................................5
3-1. Installation and dimensions....................................................................................5
3-1-1. Installation site...........................................................................................5
3-1-2. Mounting methods......................................................................................5
3-2. Wiring.....................................................................................................................7
3-2-1. Preparation.................................................................................................7
3-2-2. Cables, Terminals, glands and conduit adapter........................................8
3-3. Wiring the power supply.........................................................................................9
3-3-1. General precautions...................................................................................9
3-3-2. Access to terminal and cable entry..........................................................10
3-3-3. AC power.................................................................................................10
3-3-4. DC power.................................................................................................10
3-3-5. Grounding the housing.............................................................................10
3-3-6. Switching on the instrument.....................................................................10
3-4. Wiring the contact signals.................................................................................... 11
3-4-1. General precautions.................................................................................11
3-4-2. Contact outputs........................................................................................11
3-4-3. Contact input............................................................................................11
3-5. Wiring the mA-output signals............................................................................... 11
3-5-1. General precautions.................................................................................11
3-5-2. Analog output signals...............................................................................11
3-6. Wiring of sensors..................................................................................................12
3-6-1. Sensor cable connections using junction box (BA10) and .........................
extension cable (WF10)...........................................................................13
IM 12D08N05-01E
4th Edition: Mar. 2012 (YK)
All Rights Reserved, Copyright © 2007, Yokogawa Electric Corporation
IM 12D08N05-01E
iv
4. OPERATION OF EXAxt SC450G...............................................................................15
4-1. Main display functions..........................................................................................15
4-2. Trending graphics.................................................................................................15
4-3. Zoom in on details................................................................................................15
4-3-1. Actual mA1 ..............................................................................................16
4-3-2. Actual mA2 .............................................................................................16
4-3-3. S1/S2/S3/S4.............................................................................................16
4-3-4. C.C. (factory)............................................................................................16
4-3-5. C.C. (adjusted).........................................................................................16
4-3-6. Temp. comp 1 .........................................................................................16
4-3-7. Temp. comp 2 .........................................................................................16
4-3-8. Polarization .............................................................................................16
4-3-9. Sensor ohms ...........................................................................................16
4-3-10. Last calibrated at ...................................................................................16
4-3-11. Calibration due at...................................................................................16
4-3-12. Projected calibration at .........................................................................16
4-3-13. HART ID.................................................................................................16
4-3-14. Software revision ..................................................................................16
4-3-15. HART Device revision............................................................................17
4-3-16. Logbook.................................................................................................17
4.3.17. Trouble shooting ....................................................................................17
4-4. Information function..............................................................................................17
4-5. Setup-Calibration & commissioning.....................................................................17
4-6. Secondary- primary value display switch.............................................................17
4-7. Navigation of the menu structure.........................................................................18
5. MENU STRUCTURE COMMISSIONING....................................................................20
5-1. Configure sensor .................................................................................................20
5-2. Temperature setting..............................................................................................20
5-3. Temperature compensation..................................................................................20
5-4. Calibration settings...............................................................................................22
5-5. Concentration.......................................................................................................22
5-6. mA output setup...................................................................................................24
5-7. Contact output setup............................................................................................26
5-8. Fail........................................................................................................................28
5-9. Simulate...............................................................................................................28
5-10. Water for Injection Monitoring (USP 645 and EU 0169)....................................28
5-11. Input contacts.....................................................................................................28
5-12. Error configuration..............................................................................................30
5-13. Logbook configuration........................................................................................30
5-14. Advanced setup..................................................................................................32
5-15. Display setup......................................................................................................34
IM 12D08N05-01E
v
6. CALIBRATION............................................................................................................36
6-1. General.................................................................................................................36
6-2. Cell constant manual............................................................................................36
6-3. Cell constant automatic........................................................................................36
6-4. Air (zero) calibration ............................................................................................36
6-5. Sample calibration................................................................................................36
6-6. Temperature coefficient calibration.......................................................................36
6-7. Temperature calibration .......................................................................................37
6-8. Operation of hold function during calibration........................................................37
6-9. General comments on SC calibration...................................................................37
7. MAINTENANCE..........................................................................................................38
7-1. Periodic maintenance ..........................................................................................38
7-2. Periodic maintenance of the sensor.....................................................................38
7-3. Cleaning methods................................................................................................38
7-4. LCD adjustment....................................................................................................38
8. TROUBLESHOOTING................................................................................................40
8-1. General ................................................................................................................40
8-2. Calibration check..................................................................................................40
8-3. Polarization check................................................................................................40
8-5. Prediction of cleaning needed..............................................................................40
8-6. Poor calibration technique....................................................................................40
8-7. Error displays and actions....................................................................................40
9. QUALITY INSPECTION..............................................................................................41
10. SPARE PARTS.........................................................................................................45
APPENDICES.................................................................................................................46
Appendix 1, Temperature compensation.....................................................................46
Appendix 2, TDS readings...........................................................................................50
Appendix 3, Calibration solutions for conductivity.......................................................51
Appendix 4, Sensor selection for the EXAxt SC450G.................................................53
Appendix 5, How to enter the cell constant for the SC21¨G, SC8SG, and SC4AJ....55
Appendix 6, HART HHT (275/375) Menu structure.....................................................56
Appendix 7, Control drawing for FM approval.............................................................60
Customer Maintenance Parts List (for Style: S2).......................CMPL 12D08N05-02E
Revision Record...............................................................................................................i
IM 12D08N05-01E
vi
1
The Yokogawa EXAxt SC450G is a converter
designed for industrial process monitoring, measurement and control applications. This instruction
manual contains the information needed to install,
set up, operate and maintain the unit correctly.
This manual also includes a basic troubleshooting
guide to answer typical user questions.
Yokogawa can not be responsible for the performance of the EXAxt converter if these instructions
are not followed.
1-1. Instrument check
Upon delivery, unpack the instrument carefully and inspect it to ensure that it was not
damaged during shipment. If damage is found,
retain the original packing materials (including
the outer box) and then immediately notify the
carrier and the relevant Yokogawa sales office.
Make sure the model number on the nameplate
affixed to the top of the instrument agrees with your
order. Example of the nameplate is shown below.
KCC-REMYHQ-EEN257
CONDUCTIVITY /RESISTIVITY TRANSMITTER
SC450G
MODEL
SUFFIX
SUPPLY
OUTPUTS
AMB. TEMP.
STYLE
NO.
4-20mA(mA1: +HART)
-20 to +55°C
NI CL I, DIV 2, GROUPS ABCD
T6 FOR Ta -20 to +55°C
SEE INSTRUCTION MANUAL
TYPE 4X
SEE CONTROL DRAWING
NFM016-A9 P.1
R
C
US
Made in Japan
N200
Figure 1-1. Nameplate
Note! The nameplate will also contain
the serial number and any relevant
certification marks. Be sure to apply
correct power to the unit, as detailed
on the nameplate.
1-2. Application
The EXAxt converter is intended to be used for
continuous on-line measurement of Conductivity, Resistivity and/or Concentration in industrial
installations. The unit combines simple operation and microprocessor-based performance
with advanced self-diagnostics and enhanced
communications capability to meet the most
advanced requirements. The measurement
can be used as part of an automated process
control system. It can also be used to indicate
operating limits of a process, to monitor product
quality, or to function as a controller for a dosing/dilution system.
Sensors should normally be mounted close to
the converter���������������������������������
������������������������������������������
in order to ensure easy calibration and peak performance. If the unit must
be mounted remotely from the sensors, WF10
extension cable can be used, up to a maximum
of 50 metres (150 feet), with a BA10 junction
box, and up 10 metres standard sensor cable.
The EXAxt is delivered with a general purpose
default setting for programmable items (see
Chapter 5). While this initial configuration
allows easy start-up, the configuration should
be adjusted to suit each particular application.
An example of an adjustable item is the type
of temperature sensor used. The EXAxt can
be adjusted for a number of different types of
temperature sensors.
Details provided in this instruction manual
are sufficient to operate the EXAxt with all
Yokogawa sensor systems and a wide range
of third-party commercially available probes.
For best results, read this manual in conjunction with the corresponding sensor instruction
manual.
Yokogawa designed the EXAxt converter to
withstand industrial environments. It meets all
the CE regulatory standards. The unit meets or
exceeds stringent requirements (see section
2) without compromise, to assure the user of
continued accurate performance in even the
most demanding industrial installations.
IM 12D08N05-01E
1 INTRODUCTION AND GENERAL DESCRIPTION
1. INTRODUCTION AND GENERAL DESCRIPTION
2
2. GENERAL SPECIFICATIONS OF EXAxt SC450G
A) Inputs specifications : Two or four electrodes measurement with square wave excitation, using
max 60m (200ft) cable (WU40/WF10) and cell constants from 0.005 to
50.0 cm-1
B) Input ranges
Conductivity
: 0.000 µS/cm - 2000 mS/cm
Minimum
: 1µS (underrange 0.00 µS x c.c.)
Maximum
: 200 mS (overrange 2000 mS x c.c.)
Resistivity
Minimum
Maximum
: 0.0 Ω•cm - 1000 MΩ•cm
: 5Ω / c.c. (underrange 0.0 Ω/c.c.)
: 1MΩ / c.c. (overrange 1000 MΩ/c.c.)
Temperature
Pt1000
: -20 to 250ºC (0 - 500ºF)
Pt100
: -20 to 200ºC (0 - 400ºF)
Ni100
: -20 to 200ºC (0 - 400ºF)
NTC 8k55
: -10 to 120ºC (10 - 250ºF)
Pb36 (JIS NTC 6k) : -20 to 120ºC (0 - 250ºF)
C) Accuracy
Conductivity/resistivity : ≤ 0.5 % of reading
Temperature
: ≤ 0.3 ºC (≤ 0.4 ºC for Pt100)
Step response
: ≤ 4 sec for 90% (for a 2 decade step)
Note on performance specifications
The following tolerance is added to above performance.
mA output tolerance: ±0.02 mA of “4-20 mA“
D) Transmission signals
General
: Two isolated outputs of 4-20 mA. DC with common negative
Maximum load 600Ω. Bi-directional HART® digital communication,
superimposed on mA1 (4-20 mA) signal
Output function
: Linear or non-linear 21-step table for Conductivity/Resistivity,
concentration or temperature
Control function
: PID control
Burnout function : Burn up (21.0 mA) or burn down (3.6 mA) to signal failure acc. NAMUR
NE43
: Adjustable damping
: Expire time
Hold
: The mA-outputs are frozen to the last/fixed value during calibration/
commissioning
E) Contact outputs
General
: Four SPDT relay contacts with display indicators
Switch capacity
: Maximum values 100 VA, 250 VAC, 5 Amps. (*)
Maximum values 50 Watts, 250 VDC, 5 Amps. (*)
Status
: High/Low process alarms, selected from conductivity, resistvity,
concentration or temperature. Configurable delay time and hysteresis
: PID duty cycle or pulsed frequency control
: FAIL alarm
Control function
: On / Off
: Adjustable damping
: Expire time
IM 12D08N05-01E
3
IM 12D08N05-01E
2 GENERAL SPECIFICATIONS
Hold
: Contact can be used to signal the HOLD situation.
Fail safe
: Contact S4 is programmed as a fail-safe contact
(*)Note: When contact output current is more than 4 Amps, ambient temperature should be less
than 40 ºC.
F) Contact input
: Remote range switching to 10 times the programmed range.
Contact open
: If impedance > 100 kΩ: Range 1
(“Programmed range for mA1 output" is "Range 1.”)
Contact closed
: If impedance < 10 Ω: 10 x Range 1
G) Temperature compensation
Function
: Automatic or manual, for temperature ranges mentioned under B
(inputs).
Reference temp. : programmable from 0 to 100 ºC or 30 - 210 ºF (default 25 ºC).
H) Compensation algorithm
: According IEC 60746-3 NaCl tables (default).
Two independent user programmable temperature coefficients, from 0%
to 3.5% per ºC (ºF) by adjustment or calibration.
Matrix compensation: With conductivity function of concentration and temperature. Choice out
of 13 preprogrammed matrixes and 2 100-points user-programmable
matrixes.
I) Calibration
: Semi-automatic calibration using pre-configured OIML (KCl) buffer
tables, with automatic stability check. Manual adjustment to grab sample
J) Logbook
: Software record of important events and diagnostic data readily available
in the display or through HART®.
K) Display
: Graphical Quarter VGA (320 x 240 pixels) LCD with LED backlight and
touchscreen. Plain language messages in English, German, French,
Spanish, Italian, Swedish, Portuguese and Japanese.
L) Shipping details
Package size
: 290 x 300 x 290 mm (L x W x D) (11.5 x 11.8 x 11.5 inch)
Package weight
: app. 2.5 kg (5.5 lbs)
Converter weight : app. 1.5 kg
M) Housing
: Cast aluminum housing with chemically resistant coating;
Polycarbonate cover with Polycarbonate flexible window
:
Protection IP66 / NEMA 4X / CSA Type 3S
Colour
: Silver grey
SC450G-A(D)-A
: IP66 cable glands are supplied with the unit
SC450G-A(D)-U : NEMA 4X close up plugs are mounted in the unused cable entry holes
and can be replaced by conduit fittings as required
Pipe, Panel or Wall mounting using optional hardware
Optional conduit adapter
G1/2, 1/2NPT or M20 female
N) Power supply
SC450G-A
: Ratings; 100-240 V AC Acceptable range; 90 to 264 V AC
Ratings; 50/60 Hz
Acceptable range; 50 Hz ±5%, 60 Hz ±5%
Power Consumption; 15 VA
SC450G-D
: Ratings; 12-24 V DC
Acceptable range; 10.8 to 26.4 V DC
Power Consumption; 10 W
4
O) Safety and EMC conforming standards
,
Safety
: EN 61010-1
CSA C22.2 No.61010-1
UL 61010-1
FM3611 Class I, Div.2, Group ABCD,T6 for Ta -20 to 55ºC
EMC
: EN 61326-1 Class A, Table 2 (For use in industrial locations)
EN 61326-2-3
EN 61000-3-2 Class A
EN 61000-3-3
Korea Electromagnetic Conformity Standard
Installation altitude: 2000 m or less
Category based on IEC 61010: II (Note)
Pollution degree based on IEC 61010: 2 (Note)
Note: Installation category, called over-voltage category, specifies impulse withstand voltage.
Category II is for electrical equipment.
Pollution degree indicates the degree of existence of solid, liquid, gas or other
inclusions which may reduce dielectric strength. Degree 2 is the normal indoor
environment.
CAUTION
This instrument is a Class A product, and it is designed for use in the industrial environment.
Please use this instrument in the industrial environment only.
P) Environment and operational conditions
Ambient temperature : -20 to +55 ºC (-5 - 130 ºF)
Storage temperature : -30 to +70 ºC (-20 - 160 ºF)
Humidity
: 10 to 90% RH at 40 ºC (100 ºF) (non-condensing)
Data protection
: EEPROM for configuration data and logbook. Lithium cell for clock
Watchdog timer
: Checks microprocessor
Power down
: Reset to measurement
Automatic safeguard : Auto return to measuring mode when touchscreen is untouched for 10
min.
Model and Suffix Codes
[Style: S2]
Model
Option code
Description
SC450G
-----------------
-----------------
Conductivity/Resistivity Converter
Power
-A
-D
---------------------------------
AC version (100…240 VAC)
DC version (12…24 VDC)
---------------------------------
General purpose version
FM version
Type
-A
-U
Mounting Hardware
Hood
Conduit adapter
Tag Plate
/UM
/U
/PM
/H5
/AFTG
/ANSI
/AM20
/SCT
Universal mounting kit (panel, pipe, wall)
Pipe and wall mounting hardware (*2)
Panel mounting hardware (*2)
Awning hood (stainless steel) (*2)
G1/2 (*2)
1/2NPT (*2)
M20 (*2)
Stainless steel tag plate (*1)
*1 If the tag number is predefined with the purchase, Yokogawa will inscript the tag plate with the specified tag number, and
program the tag number in the converter.
*2 Option codes /U, /PM, /H5, /AFTG, /ANSI and /AM20 are not specified for FM version (-U).
IM 12D08N05-01E
5
3. INSTALLATION AND WIRING
3-1-2. Mounting methods
Refer to figures 3-2 and 3-3. Note that the
EXAxt converter has universal mounting capabilities:
• Panel mounting using optional brackets
• Surface mounting on a plate (using bolts
from the back)
• Wall mounting on a bracket (for example,
on a solid wall)
• Pipe mounting using a bracket on a
horizontal or vertical pipe
Size nominal 50A
Unit: mm (inch)
Hood (Option code: /H5, some cutout on the left side cover)
184(7.2")
220(8.66")
144(5.67")
72(2.8")
24.5(1")
144(5.67")
20
(0.79")
M20 cable gland
(When shipped, not installed)
20
(0.79")
121.5(4.78")
A
D
B
C
E
F
Adapter for Coduit Work (optional)
(option code : / AFTG, / ANSI, / AM20)
A : For output signal
Adapter
B : For contact input
C : For sensor cable
D : For contact output (S1 and S2)
E : For contact output (S3 and S4)
F : For power supply
Grounding terminal (M4 screw)
Approx.
49
(1.93") 55(2.2")
G1/2 screw (/AFTG), 1/2 NPT screw (/ANSI)
M20 screw (/AM20)
Figure 3-1. Housing dimensions and layout of glands
IM 12D08N05-01E
3 INSTALLATION AND WIRING
3-1. Installation and dimensions
3-1-1. Installation site
The EXAxt 450 converter is weatherproof and
can be installed inside or outside. It should,
however, be installed as close as possible to
the sensor to avoid long cable runs between
sensor and converter. In any case, the cable
length should not exceed 60 metres (197 feet).
Select an installation site where:
• Mechanical vibrations and shocks are
negligible
• No relay/power switches are in the direct
environment
• Access is possible to the cable glands
(see figure 3-1)
• The converter is not mounted in direct
sunlight or severe weather conditions
• Maintenance procedures are possible
(avoiding corrosive environments)
The ambient temperature and humidity of the
installation environment must be within the
limits of the instrument specifications. (See
chapter 2).
6
Unit: mm (inch)
+1
138 0
(5.43")
min.195(7.75")
min.185 (7.25")
+1
138 0
(5.43")
138(5.43")
M6
138
(5.43")
M5
M6
Figure 3-2. Option /PM: panel mounting diagram
(Note) When option code "/UM" is specified, universal pipe/wall/pannel mounting kit are supplied--same as option code "/U" and "/PM" both specified.
wall mounting
13
(0.51")
80
(3.15")
pipe mounting
(vertical)
pipe mounting
(horizontal)
2x Ф6.5
(0.26")
200
(7.87")
4x Ф10
(0.4")
35
(1.38")
15
(0.6")
70
(2.75")
154.5
(6.08")
2" ND. pipe
Figure 3-3. Option /U: wall and pipe mounting diagram
IM 12D08N05-01E
7
WARNING This connector for software must be used only by
Yokogawa’s
service
personnel.
connector
for (future)
software
input
terminal
block
potentiomerter
LCD
bracket
protective
shield
bracket
6 X M20 glands
Note: SC450G-A(D)-U
The enclosure is provided with stoppers in stead of M20 cable glands for the unused holes.
These stoppers must be removed and replaced by FM approved conduit fittings in accordance with good installation practice. Also see Appendix 7, Control drawing for FM approval.
Figure 3-4. Internal view of EXAxt wiring compartment
3-2. Wiring
3-2-1. Preparation
Refer to figure 3-4. The relay contact terminals
and power supply connections are under the
screening (shielding) plate. These should be
connected first. Connect the sensor, outputs
and HART® communication connections last.
To open the EXAxt 450 for wiring:
1. Loosen the four frontplate screws and swing
open the cover.
2. The upper terminal strip is now visible.
3. Remove the screen (shield) plate covering
the lower terminal strip.
4. Connect the power supply and contact
outputs. Use the three glands at the back for
these cables.
DANGER
• Cables that withstand temperatures of at
least 70 °C should be used for wiring.
• Wiring work should be performed to meet
IP66 or higher requirements. Tighten four
frontplate screws to 1.5 N·m torque.
WARNING
Always place the screen plate over the power
supply and contact terminals for safety reasons
and to avoid interference.
5. Put back (replace) the screen (shield) plate
over the lower terminals.
6. Connect the analog output(s), the sensor
inputs, and, if necessary, the HART® wiring
and input contact.
7. Use the front three glands for analog output,
sensor inputs, contact input and HART®
wiring (see figure 3-5).
8. Swing back the cover and secure it with the
four screws.
9. Switch on the power. Commission the
instrument as required or use the default
settings.
CAUTION
Do not turn on power with the touchcsreen
pressed, otherwise inaccurate screen
positioning will occur. If it occurs, leave the
touchscreen unpressed, turn off power then
on again. The screen positioning will be
accurate.
IM 12D08N05-01E
3 INSTALLATION AND WIRING
output
terminal
block
8
3-2-2. Cables, Terminals, glands and conduit
adapter
SC450G-A(D)-A
The SC450 is supplied with terminals suitable
for the connection of finished wires in the size
range of 0.13 to 2.5 sq.mm. ( 26 to 14 AWG).
The cable glands supplied will form a tight seal
on cables with an outside diameter of 6 to 12
mm (0.24 to 0.47 inches). Unused cable entry
holes must be sealed with cable glands including the blind plugs supplied.
SC450G-A(D)-U
The SC450 is supplied with terminals suitable
for the connection of finished wires in the size
range of 14- 26 AWG. The cable entry holes
are sealed with FM certified plugs. Prior to
cable entry the plugs can be removed with
allen key size 3/8” The cable conduit fittings
can be mounted in the holes of the housing as
required. The cable glands supplied with the
unit will give a tight seal on cables with outside
diameter of 0.24 to 0.47 inches.
High voltage section
Contact
(S1, S2)
output
cables
mA
cables
Contact
(S3, S4)
output
cables
Input
contact
Power
cable
Sensor
Cables
Suitable for cables with an outside diameter between 6 - 12 mm (0.24 - 0.47”)
Figure 3-5a. Cable glands diagram
Gland nut
Contents:
6 X Gland M20
6 X Close up plug
6 X Gland nut M20
6 X O-ring 17.12 X 2.62 EPDM 70° sh.
O-ring
NOTE: The glands must be installed properly
to meet IP66 and NEMA 4X rating.
Use close up plug for unused glands.
Gland
Close up plug
Figure 3-5b. How to install cable glands
IM 12D08N05-01E
NOTE: Moisturize O-ring before assembling.
9
Adapter for conduit work
When protect the cable with a conduit, replace the M20 cable gland with a cable gland of optional
conduit adapter, and set the adapter shown as Figure 3-5c.
Unit: mm(inch)
Nut
Packing
G1/2 screw (/AFTG), 1/2 NPT screw (/ANSI)
M20 screw (/AM20)
Adapter
Figure 3-5c. Conduit adapter
FRONT GLANDS
REAR GLANDS
Sensor
mA1
HART
output
signals
Power
Contact
output
S1
S2
Contact
output
mA2
S3
S4
Contact input
Figure 3-6. System configuration
3-3. Wiring the power supply
3-3-1. General precautions
Make sure the power supply is switched off.
Also, make sure that the power supply is correct for the specifications of the EXAxt and that
the supply agrees with the voltage specified on
the textplate.
DANGER
1. Install an external switch or circuit breaker
to the power supply of the converter.
2. Use an external switch or circuit breaker
rated 5A and conforming to IEC 60947-1
or IEC 60947-3.
3. It is recommended that the external switch
or circuit breaker be installed in the same
room as the converter.
4. The external switch or circuit breaker
should be installed within reach of the
operator and identified with marking as a
power supply switch to the converter.
5. Power lines such as power cables and
contact outputs should be fixed securely
onto a wall or construction using cable
racks, conduit tubing, nylon bands or
other appropriate ways. Accidental
removal from terminals by pulling may
result in electric shock.
Local health and safety regulations may require
an external circuit breaker to be installed. The
instrument is protected internally by a fuse. The
fuse rating is dependent on the supply to the
instrument. The 250 VAC fuses should be of
the “time-lag” type, conforming to IEC127.
IM 12D08N05-01E
3 INSTALLATION AND WIRING
Approx.
55(2.2") 49
(1.93")
10
3-3-3. AC power
Connect terminal L to the phase line of the AC
power and terminal N to the zero line. See figure 3-8 for the power ground. This is separated
from input ground by a galvanic isolation.
WARNING
Fuse replacement should be performed only by
a qualified service personnel.
See Sec.7. MAINTENANCE, Fuse
3-3-4. DC power
Connect terminal 1 to the positive outlet and
terminal 2 to the negative outlet.
This is separated from input ground by a
galvanic isolation. The size of conductors
should be at least 1.25 mm2. The overall cable
diameter should be between 6 & 12 mm.
Fuse ratings:
Power supply
Fuse type
12-24VDC, 10W max
2A/250V, Slow
100-240VAC, 15VA max 0.5A/250V, Slow
3-3-2. Access to terminal and cable entry
Terminals 1 and 2 are used for the power supply. Guide the power cables through the gland
closest to the power supply terminals. The terminals will accept wires of 2.5 mm2 (14 AWG).
Always use cable finishings if possible.
2
1
-
2
+
1
N
POWER
L
POWER
100-240 VAC/15 VA/ 50/60Hz
FUSE: 500 mA/250 VAC/T
12-24 V
/10 W
FUSE: 2A/250 VAC/T
AC
DC
32
31
NC
33
C
42
NO
NC
S1
63
SHLD
66
-
41
C
NO
S2
65
+
mA2
62
61
+
mA1
22
250V / 5A
52
AC / DC
100VA / 50W NC
43
CONTACTS
21
11
+
(+HART)
12
-
51
C
53
72
NO
NO
S3
71
C
73
NC
S4 (fail-safe)
13
14
15
16
INNER OUTER INNER OUTER
TEMP
CONTACT
SENSOR(S)
mA OUTPUTS
REFER TO INSTRUCTION MANUAL FOR CONNECTIONS
ELECTRODE
SC
Figure 3-7. Input and output connections
3-3-5. Grounding the housing
For the safety of the user and to protect the
instrument against interference, the housing
must always be connected to ground. This has
to be done by a large area conductor. This
cable can be fixed to the rear of the housing or
by using the internal ground connections using
a braided wire cable. See figure 3-8.
DANGER
The minimum cross sectional area of the protective grounding wire should be 0.75 mm2.
IM 12D08N05-01E
3-3-6. Switching on the instrument
After all connections are made and checked,
the power can be switched on from the power
supply. Make sure the LCD display comes on.
After a brief interval, the display will change to
the measured value. If errors are displayed or
a valid measured value is not shown, consult
the troubleshooting section (Chapter 8) before
calling Yokogawa.
11
(M4 screw)
2
1
L
POWER
100-240 VAC/15 VA/ 50/60Hz
FUSE: 500 mA/250 VAC/T
AC
(M4 screw)
Figure 3-8-a. External grounding
3-4. Wiring the contact signals
3-4-1. General precautions
The contact output signals consist of voltage-free relay contacts for switching electrical
appliances (SPDT). They can also be used as
digital outputs to signal processing equipment
(such as a controller or PLC). It is possible to
use multi-core cables for the contact in and
output signals and shielded multi-core cable for
the analog signals.
3-4-2. Contact outputs.
The EXAxt 450 unit’s four contacts (switches)
that can be wired and configured to suit user
requirements. Contact S4 is programmed as
a fail-safe contact. Please refer to section 5-7,
Contact output setup for functionality description.
Alarm (limits monitoring)
Contacts configured as “ALARM” can be energized when limits are crossed.
Fail
Contacts configured as “FAIL” will be energized
when a fail situation occurs. Some fail situations are automatically signaled by the internal
diagnostics (electronics) of the converter.
Others can be configured by the user (see
section 5-12 Error Configuration). By pressing
the “INFO” button on the main screen the user
is given an explanation as well as a remedy for
the current fail situation.
Always connect the fail contact to an alarm
device such as a warning light, alarm bell or
displayed on an annunciator.
Figure 3-8-b. Internal grounding
“ALARM” Contact “FAIL” Contact
Power Off
NC
NC
Power On
NC
NC
Alarm
NO
NC
Fail
NC
NO
Fail and Alarm
NC*
NO
HOLD
NC
NC
* W
hen a fail situation occurs which is related
to the parameter associated with the contact
(Conductivity, Concentration or temperature)
the contact will go to NC. When the fail
situation is not related to the parameter
associated with the contact the contact will
remain in the state it is currently in.
3-4-3. Contact input
It is necessary to use screening/shielding on
the input signal cables. Terminal 63 is used to
connect the shielding.
3-5. Wiring the mA-output signals
3-5-1. General precautions
The analog output signals of the EXAxt
transmit low power standard industry signals to
peripherals like control systems or strip-chart
recorders (Figure 3-6).
3-5-2. Analog output signals
The output signals consist of active current
signals of 4-20 mA. The maximum load can be
600 ohms on each.
It is necessary to use screening/shielding on
the output signal cables. Terminal 63 is used to
connect the shielding.
IM 12D08N05-01E
3 INSTALLATION AND WIRING
N
12
3-6. Wiring of sensors
General precautions
Generally, signals from sensors are at low voltage and current level. Thus a lot of care must
be taken to avoid interference. Before connecting sensor cables to the converter make sure
that following conditions are met:
– the sensor cables are not mounted in
tracks together with high voltage and or
power switching cables
– only standard sensor cable or extension
cable is used
– the converter is mounted within the distance of the sensor cables (max. 10 m) +
up to 60m WF10 extension cable.
– the setup is kept flexible at the sensors end
for easy insertion and retraction of the sensor in the fitting.
Figure 3-9. Sensor wiring diagrams
IM 12D08N05-01E
Sensor wiring
Refer to figure 3-9, which includes drawings
that outline sensor wiring.
The EXAxt can be used with a wide range of
sensor types. The sensor system fall into two
categories, the ones that use fixed cables and
the ones with separate cables.
To connect sensors with fixed cables, simply
match the terminal numbers in the instrument
with the identification numbers on the cable
ends.
13
11
12 13 14 15 16
11
12 13 14 15 16
t
t
3-6-1. Sensor cable connections using junction box (BA10) and extension cable
(WF10)
Where a convenient installation is not possible
using the standard cables between sensors
and converter, a junction box and extension
cable may be used. The Yokogawa BA10
16 15
14 14
17
13 12 11
11 12 13 17
15 16
14 14
BA10
junction box and the WF10 extension cable
should be used. These items are manufactured
to a very high standard and are necessary to
ensure that the specifications of the system
can be met. The total cable length should not
exceed 60 metres (e.g. 5 m fixed cable and 55
m extension cable).
WF10
15 Core 16 Screen
White Co-axial cable
14 Overall Screen
13 Core 17 Screen
Brown Co-axial Cable
11 Red
12 Blue
EXA TRANSMITTER
/ CONVERTER
11
11
12
12
13
13
14
14
15
15
16
16
17
17
12 (blue)
14 (overall screen)
13 (core)
Co-axial cable 15 (core)
(white)
16 (screen)
17 (screen)
Co-axial cable
(brown)
11 (red)
Figure 3-10.
Connection of WF10 extension cable and BA10 junction box
IM 12D08N05-01E
3 INSTALLATION AND WIRING
Other sensor systems
To connect other sensor systems, follow the general pattern of the terminal connections as listed
below:
11 and 12: Always used for temperature compensation resistor input.
13 and 14: Normally used for the outer electrode
15 and 16: Used for inner electrode
In case a 4-electrode measuring system will be used, 14 and 16 should be used for the current
electrodes. Please ensure that shielded cabling will be used.
In below figure this is shown in a schematic way.
14
Extension cable may be purchased in bulk
quantities or in pre-finished lengths. In the case
of bulk quantities cut to length, then it is necessary to terminate the cable as shown below.
Termination procedure for WF10 cable.
1. Slide 3 cm of heat shrink tube (9 x 1.5)
over the cable end to be terminated.
2.
Strip 9 cm of the outer (black) insulating
material, taking care not to cut or damage
internal cores.
3. Remove loose copper screening, and cut
off the cotton packing threads as short as
possible.
3 cm
heat shrink
4.
Strip insulation from the last 3 cm of the
brown, and the white coaxial cores.
5. Extract the coaxial cores from the
braid, and trim off the black (low-noise)
screening material as short as possible.
6. Insulate the overall screen and the 2
coaxial screens with suitable plastic
tubing.
7.
Strip and terminate all ends with suitable
(crimp) terminals and identify with
numbers as shown.
8.
Finally shrink the overall heat shrink tube
into position.
9 cm
remove insulation
Figure 3-11.a.
3 cm
cotton threads
Figure 3-11.b.
11
12
17
15
16
13
14
Figure 3-11.c.
IM 12D08N05-01E
15
4. OPERATION OF EXAxt SC450G
4-1. Main display functions
Tag:EXAxt SC450
200.0
25.0
16.64
Go to trend screen
+
μS/cm
Go to zoom screen
Go to status screen
4-3. Zoom in on details
This button gives access to the diagnostic information of the analyzer. The following messages
will appear under normal (default) conditions:
Zoom in on Details
- Home key back to
mainscreen.
- One level up.
Go to maintenance screen
Figure 4-1. Main Display
SC 450
Next
Next
Next
- Enter selected data
or choice.
First zoom screen
gives you inside
into the parameters
involving current
measurement. All
following zoom
screens give
additional information
about the device and
lead to logbook data.
Figure 4-3. Detail screen
Maximum
Live reading
120.0
SC
Next
90.0
Minimum
60.0
30.0
109.3 µS/cm
T
Maximum
Average
Minimum
Figure 4-2. Trend screen
IM 12D08N05-01E
OPERATION OF EXAxt SC450G
4-2. Trending graphics
Pressing the button changes the display
into a graphical mode in which the average
measured value is shown on a time scale. The
“Live” value is also digitally displayed in a text
box. The time scale ( X-axis) and the primary
value scale (Y-axis) are set in the “DISPLAY
SETUP” menu. The full screen displays a trend
of 51 points that represent the average of the
selected time interval. The analyzer samples
the measurement every second. The trending
graphic also shows the maximum and minimum
measured value in that interval.
For example if the time scale is set to 4 hours,
then the trend is shown for 4 hours prior to the
actual measurement. Each point on the trend
line represents the average over 4*60*60/51=
282 measurements (seconds).
- Scroll choices (grey
means deactivated).
4
A heart “ ” mark is lit on the right-upper corner of the
screen when HART communication is active. An “Χ”
mark is lit when HART communication is abnormal.
Nothing appears when HART communication is not
used.
Note that the “Χ” mark may appear due to output
signal noise or the like even when HART communication is not used. There is not problem when HART
communication is not used. Continue operation while
ignoring the mark.
16
4-3-1. Actual mA1 = the current output in mA of
the first current output, which is defined as mA1.
The range and function of this mA output can be
set in:
Routing: Commissioning >> Output setup
>> mA1
4-3-2. Actual mA2 = the current output in mA of
the second current output, which is defined as
mA2. The range and function of this mA output
can be set in:
Routing: Commissioning >> Output setup
>> mA2
4-3-3. S1/S2/S3/S4 = the current state of
contacts 1 to 4. The functions and settings of the
contacts can be set in:
Routing: Commissioning >> Output setup
>> S1/S2/S3/S4
4-3-4. C.C. (factory) = the nominal cell constant
as determined by the factory calibration during
production. This value is set during commissioning, and is found on the nameplate of the sensor
or the calibration certificate.
Routing: Commissioning >> Measurement
setup >> Configure sensor
4-3-5. C.C. (adjusted) = the calibrated cell
constant. When the cell constant of the system is
adjusted on-line by grab sample or by calibrated
solution technique, the new cell constant is
recorded here. This value should not deviate
greatly from the original factory calibration. In
the event that there is a significant discrepancy
seen between this reading and the C.C. (factory)
value, the sensor should be checked for damage
and cleanliness.
Routing is via the “Calibration” menu.
4-3-6. Temp. comp 1 = the chosen temperature
compensation method for the primary measurement.
Routing: Commissioning >> Measurement
setup >> Temp.compensation
IM 12D08N05-01E
4-3-7. Temp. comp 2 = the chosen temperature
compensation method for the secondary measurement.
Note: This does not imply two separate measurements. There is the possibility to set two
separate compensation methods so that two
different stages of the same process can be
monitored accurately. An example is process/
cleaning fluid interface.
Routing: Commissioning >> Measurement
setup >> Temp.compensation
4-3-8. Polarization = the polarization is measured by the input circuitry. Monitoring this figure
gives a guide to progressive fouling of the sensor.
4-3-9. Sensor ohms = the input measurement
as an uncompensated resistance value.
4-3-10. Last calibrated at = the date of the last
calibration
4-3-11. Calibration due at = the date scheduled
for the next calibration. This field is determined by
the calibration interval.
Routing: Commissioning >> Measurement
setup >> Calibration settings
4-3-12. Projected calibration at =
a diagnostic output, showing a time frame when
the unit should next be maintained according to
the sophisticated self-diagnostic tools built into
the EXAxt software (for example >12 months,
3-6 months or 0-1 month).
The analyzer checks the rate of polarization
every 24 hours. If a clear increase of polarization is observed, the user is notified when a
next calibration should take place. Prior to
calibration the sensor should be well cleaned
and rinsed.
4-3-13. HART ID = a part of the HART device ID
(descriptor)
4-3-14. Software revision = the revision level of
the software in the instrument.
17
4-3-15. HART Device revision
Sometimes the firmware of a device is
updated in a way that the communication file
(HART DD) need revision too. In this case the
revision level is increased by one. The revision
level of the HART DD must match the revision
level of the Firmware. The revision level is
expressed by the first two characters of the
filename. The following files should be used
when the HART Device revision level is 2.
(0201.aot, 0201.fms, 0201.imp, 0201.sym)
Sensor will give all historical information on
parameter settings concerning the sensor(s).
The events logged in this logbook are user
definable. This is done in: Commissioning >>
Configure Logbook >> Sensor Logbook.
Predictive maintenance. If the sensor diagnostics of the EXAxt are enabled, the diagnostics are saved into this logbook.
For the EXAxt SC450G, the polarization (due
to fouling) is stored once a day. This information can be used for (predictive) maintenance
schedules as the polarization is a measure of
fouling and the sensor should be kept clean for
best results.
Settings wil give all history information on parameter settings concerning the analog outputs
(mA1/mA2) and contact (S1…S4). This logbook
is useful to trace back differences in performance due to changed settings. The events
logged in this logbook are user definable. This
is done in: Commissioning >> Configure
Logbook >> Settings Logbook – mA and/or
Settings Logbook – contact
Each HMI screen can contain up to 5 events.
As each logbook can contain 50 events in total,
one can access previous events by selecting
another page 1 to 10.
4.3.17. Trouble shooting
If you contact the local sales/ service organization the serial number and software revision is
necessary information. Without that information
it is impossible to help you. It is also very useful
to report all the information that you find on the
zoom-in display.
4-4. Information function
In this field an information sign , a warning
sign
or a fail sign
can appear. Pushing
this button, the user gets detailed information
about the status of the sensor or the instrument
if applicable.
See troubleshooting (chapter 8) for further
details.
OPERATION OF EXAxt SC450G
Calibration will give information of previous
calibrations. This logbook is useful as one now
can
1) Monitor the sensor performance over time.
2) Monitor the sensor(s) lifetime.
S1/S2/S3/S4 shows all (dynamic) events concerning the contacts.
4
4-3-16. Logbook
The EXAxt contains several logbooks to store
historical information on events, changed
settings and calibrations. The logbooks have
been categorized to simplify the retrieval of this
information.
mA1/mA2 shows all (dynamic) events concerning the analog outputs
4-5. Setup-Calibration & commissioning
By pressing the setup
key, you get access
to the operating system of the converter based
on menus and submenus.
Browse through the list using the
key
till you find the required menu and press the
key to enter this menu.
It is also possible to press on the
or
symbol found beside the menu item.
4-6. Secondary- primary value display switch
25.0
Pressing on this text block automatically
switches the secondary value to the main
display (Large font size).
IM 12D08N05-01E
18
4-7. Navigation of the menu structure
Instrument in HOLD
“RETURN KEY” exit to previous display
Main display
Primary setup display
Commisioning menu display
“HOME KEY” return to main display
IM 12D08N05-01E
19
4
OPERATION OF EXAxt SC450G
IM 12D08N05-01E
20
5. MENU STRUCTURE COMMISSIONING
Measurement setup
Main parameter
Choose the required parameter, either conductivity or resistivity. If the main parameter is
changed the instrument will reset main display
settings, units and recalculate several values.
The menu structure will change accordingly.
5-1. Configure sensor
Sensor type
Choose the sensor type used. Normally
conductivity and/or resistivity measurements
are done with 2-electrode type sensors. At
high conductivity ranges, polarization of the
electrodes may cause an error in conductivity
measurement. For this reason 4-electrode type
sensors may be necessary.
Measuring unit /cm /m
Either /cm or /m can be chosen here. The process values will be expressed in S/cm or S/m
respectively, (Ω.cm or Ω.m in resistivity mode).
Cell constant (factory)
Cell constant given by factory calibration.
Usually given on a label on the sensor or the
calibration certificate. Only change this value
in case a new sensor is used. By changing this
value the actual cell constant is also changed.
Measure
Process values to be measured can be selected to suit the user’s preference.: Conductivity only, Concentration only or one of both
Conductivity and Concentration.
Note: this choice is not available in
Resistivity mode.
5-2. Temperature setting
Temperature Element
Selection of the temperature sensor used for
compensation. The default selection is the
Pt1000 Ohm sensor, which gives excellent
precision with the two wire connections used.
The other options give the flexibility to use a
very wide range of other conductivity/resistivity
sensors.
IM 12D08N05-01E
Temperature Unit
Celcius or Fahrenheit temperature scales can
be selected to suit the user’s preference.
When the unit is changed all temperature
related parameters and settings will be recalculated.
5-3. Temperature compensation
Compensation
Two types of methods can be used here. Automatic for use of temperature element. Select
one of the Temperature elements used. The
other is a manual set temperature. The manual
temperature that represents the process temperature must be set here.
Reference Temperature
Choose a temperature to which the measured conductivity (or resistivity) value must be
compensated. Normally 25°C (77ºF) is used,
therefore this temperature is chosen as the
default value.
Method
TC In addition to the temperature coefficient
calibration routine it is possible to adjust the
compensation factor directly. If the compensation factor of the sample liquid is known from
laboratory experiments or has been previously
determined, it can be introduced here.
Adjust the value between 0.00 to 3.50 % per
°C. In combination with reference temperature
a linear compensation function is obtained,
suitable for all kinds of chemical solutions.
NaCl Temperature compensation according
NaCl curve. See appendix 1 for values.
Matrix The EXAxt is equipped with a matrix type algorithm for accurate temperature
compensation in various applications. Select
the range as close as possible to the actual
temperature/concentration range. The EXAxt
will compensate by interpolation. If user defined
1 or user defined 2 is selected, the temperature
compensation range for the adjustable matrix
must be defined.
See Appendix 1 for matrix interpolation.
Note! Extra information on temperature
compensation is given in Appendix 1.
21
Conductivity only
Resistivity only
Conduct.+Concentr
Concentration only
Measurement setup
Measure Conductivity only
Configure sensor
Temperature settings
Temp. Compensation
Calibration settings
Concentration
Enter
5
Menu
Parameter Default
Range
values
min.
Configure Sensor
Cell constant
0.1 cm-1
0.005 cm-1
Temp. Comp.
Manual Comp.
Temp. Coef
Temp. Coef
Reference Temp.
Manual Temp.
T.C.methods 1
T.C.methods 2
25ºC, 77ºF
25ºC, 77ºF
2.10%/ºC, 1.17%/ºF
2.10%/ºC, 1.17%/ºF
0ºC, 32.0ºF
-20ºC, -3.0ºF
-10%/ºC, -0%/ºF
0%/ºC, 0%/ºF
max.
50 cm-1
100ºC, 211ºF
250ºC, 482ºF
3.5%/ºC, 2%/ºF
3.5%/ºC, 2%/ºF
IM 12D08N05-01E
MENU STRUCTURE COMMISSIONING
Note!
“Main parameter” and/or “Measure” determines the rest of the HMI menu structure
22
5-4. Calibration settings
Air adjust limit
To avoid cable influences on the measurement,
a “zero” calibration with a dry sensor may be
done. If a connection box (BA10) and extension
cable (WF10) are being used, “zero” calibration should be done including this connection
equipment.
When using a 4-electrode sensor additional
connections are required. Temporarily Interconnect terminals 13 & 14 with each other and 15
& 16 with each other before making the adjustment. This is necessary to eliminate the capacitive influence of the cables. The links should be
removed after this step is completed.
As the calibration is performed in air the
resistivity is infinite (open connection). Higher
conductivity values than the air adjust limit
indicate the cell is not in air or is still wet. To
prevent wrong air calibrations a limit must be
given here.
c.c. high limit
High limit of the cell constant expressed in %
of nominal value. During calibration this value
is used to check if the calibrated cell constant
remains within reasonable limits.
c.c. low limit
Low limit of the cell constant expressed in %
of nominal value. During calibration this value
is used to check if the calibrated cell constant
remains within reasonable limits.
Stabilization time
During calibration the stability of the measurement is constantly monitored. When the value
is within a bandwidth of 1% over a period of the
stabilization time, the calibration is considered
stable and the calibration may be completed.
Calibration Interval
A user defined interval in which a new calibration should take place. If the interval is exceeded the instrument will give a warning or a
fail (user definable in error configuration 2/3)
IM 12D08N05-01E
5-5. Concentration
Concentration has a direct relation with the
conductivity value at reference temperature.
This relation is built in every matrix which are
used for temperature compensation. These can
be found in:
Commissioning >> Measurement setup >>
Temp. compensation >> Method
By selecting one of the matrices for temperature compensation directly gives the
concentration value on the main display. If
another temperature compensation method is
chosen (NaCl or T.C.), the relation between the
conductivity at reference temperature and the
concentration is obtained from the “Concentration table”.
Additional table
This 21x2 user defined concentration table is
used to come to more accurate concentration
values compared to the temperature compensation matrix. Enabling this additional table
overrules the concentration values obtained
from the matrix (if used).
Unit for table
The way the concentration values are presented to the user. Changing the unit will not
result in a re-calculation of the table.
23
Measurement setup
Measure Conductivity only
Configure sensor
Temperature settings
Temp. Compensation
Calibration settings
Concentration
Enter
Concentration measurement is only possible if “measure” in the
“Configure sensor” menu is set to “conductivity + concentration”
or “Concentration only”.
5
MENU STRUCTURE COMMISSIONING
Menu
Parameter Default
Range
values
min.
Calibration
Air adjust
10.00 µS
0 µS
c.c. high
120%
100%
c.c. low
80%
80%
Stabilization time
5 s
2 s
Calib. interval
250 days
1 day
Concentr. Table
Table
See appendix
max.
20 µS
120%
100%
30 s
250 days
IM 12D08N05-01E
24
5-6. mA output setup
The general procedure is to first define the
function (control, output, simulate, off) of the
output and second the process parameter associated to the output.
Available process parameters depend on the
selected “main parameter” and “measure”.
Off
: When an output is set off the
output is not used and will give an
output of 4 mA
Control : A selection of P- PI- or PID control
Manual : Static output required to maintain
reset
equilibrium state with setpoint
Direction : Direct
If the process variable is too high
relative to the SP, the output of the
controller is increased
(direct action).
: Reverse
If the process variable is too high
relative to the SP, the output of the
controller is decreased
(reverse action).
Output : Linear or non linear table output.
The table function allows the confi
guration of an output curve by 21
steps (5% intervals). In the main
menu concentration can be selec
ted to set the concentration range.
Simulate : Percentage of output span.
Normal span of outputs are limited
from 3.8 to 20.5 mA
Fail safe : Contact S4 is programmed as a failsafe contact.
Burn Low or High will give an output of 3.6
resp. 21 mA in case of Fail situation.
Note! When leaving Commissioning, Hold
remains active until switched off
manually. This is to avoid inappropriate
actions while setting up the
measurement
Proportional control
Proportional Control action produces an output
signal that is proportional to the difference
between the Setpoint and the PV (deviation or
error). Proportional control amplifies the error to
motivate the process value towards the desired
setpoint. The output signal is represented as a
percentage of output (0-100%).
IM 12D08N05-01E
Proportional control will reduce but not
eliminate the steady state error. Therefore,
proportional Control action includes a Manual
Reset. The manual reset (percentage of output)
is used to eliminate the steady state error.
Note! Any changes (disturbances) in the
process will re-introduce a steady
state error. Proportional control can
also produce excessive overshoot and
oscillation. Too much gain may result
in an unstable- or oscillating process.
Too little gain results in a sustained
steady state error. Gain = 1/Range.
[PV units]
Integral Control
Integral control is used to eliminate the steady
state error and any future process changes.
It will accumulate setpoint and process (load)
changes by continuing to adjust the output
until the error is eliminated. Small values of
integral term (I-time in seconds) provide quick
compensation, but increase overshoot. Usually,
the integral term is set to a maximum value that
provides a compromise between the three system characteristics of: overshoot, settling time,
and the time necessary to cancel the effects of
static loading (process changes). The integral
term is provided with an anti windup function.
When the output of PI portion of the controller
is outside the control range (less than -5% or
greater than 105%), the I-part is frozen.
Derivative control
The control acts on the slope (rate of change)
of the process value, thereby minimizing
overshoot. It provides “rate” feedback, resulting
in more damping. High derivative gains can
increase the rizing time and settling time. It is
difficult to realize in practice because differentiation leads to “noisy” signals.
z
Controller
SP
+
e
-
e
Range
+
+
1
∫e dt
Ti
+
Actuator
-
dPV
Td
dt
Figure 5-1. Control diagram
+
+
Process
Process
PV
25
5
MENU STRUCTURE COMMISSIONING
mA2 similar structure to mA1
Menu
Parameter Default
Range
values
min. max.
mA1 (control)
mA1 (output)
mA1 (simulate)
Expire time
Damping time
Simulation perc.
0.0 sec.
0.0 sec.
50%
0 sec.
0 sec.
0%
1800 sec.
3600 sec.
100%
PID-control mA1
PID-control mA2
PID-control mA1
PID-control mA2
PID-control mA1
PID-control
PID-control
Setpoint
Setpoint
Range
Range
Manual Reset
I-time
D-time
250.0 µS/cm
25ºC/ºF
50.00 µS/cm
10ºC/ºF
0%
3600 sec.
0 sec.
- inf
- inf
- inf
- inf
0%
1 sec.
0 sec.
+ inf
+ inf
+ inf
+ inf
100%
3600 sec.
60 sec.
Linear mA1
Linear mA2
Linear mA1
Linear mA2
Table
0% Value
100% value
Table mA1
0 S/cm
0ºC/ºF
500.0 μS/cm
100ºC/ºF
see appendix
-
+
+
+
+
+
inf
inf
inf
inf
inf
inf
inf
inf
inf
inf
IM 12D08N05-01E
26
Damping time
The response to a step input change reaches
approximately 90 percent of its final value
within the damping time.
applies. For calibration, it is up to the user if
HOLD is enabled or not.
Setpoint
Hys.
SC
Expire time
If the output is over 100% for longer than the
expire time, the output will return to 0%.
range
100%
off
Direct
manual
reset
0%
Delay time
set process
point value
range
100%
0%
Duty cycle
50%
50
5-7. Contact output setup
S1/S2/S3/S4
Each Switch (contact) can have the following
functions.
1. Control : A selection of P- PI- or PID control
2. Alarm : Low or high value Limits monitoring
3. Hold : A hold contact is energised when
the instrument is in HOLD
4. Fail : S4 is set as fail-safe contact.
5. Simulate : To test the operation of the contact,
simulate can be used. The contact
can be switched on or off or a
percentage of duty cycle can be
entered (DC period time)
6. Off
: Switch is not used.
7. USP : USP/EU limits for WFI
power on
contact
activated
S1, S2, S3
S4
Above table shows contact output status between common to NO.
Configure hold
Hold is the procedure to set the outputs to
a known state when going into commissioning. During commissioning HOLD is always
enabled, outputs will have a fixed or last value.
During calibration the same HOLD function
IM 12D08N05-01E
t (sec)
toff > 0.1 sec
Figure 5-2. Direct/Reverse action
power on
normal opened
Delay time
% controller output
100
set process
point value
power down
off
Figure 5-3. Alarm contact (on/off control)
Reverse
manual
reset
on
50%
ton
toff
Duty cycle
ton > 0.1 sec
0
Range
Duty cycle
Figure 5-4. Duty cycle control
% controller output
0.3 s
100
Maximum pulse frequency
0.3 s
50
0
50% pulse frequency
No pulses
Range
Figure 5-5. Pulse frequency control
Lifetime contacts
One should note that the lifetime of the contacts is limited (106) When these contacts are
used for control (pulse frequency or duty cycle
with small interval times) the lifetime of these
contact should be observed. On/Off control is
preferred over Pulse/duty cycle.
27
PI-control S1
Setpoint
Range
(proportional)
Direction
I-time
250.0μS/cm
50.00μS/cm
Reverse
3600 s
Enter
Control
Alarm
Hold
USP
Fail
Simulate
Off
5
MENU STRUCTURE COMMISSIONING
S2, S3, S4 Similar structure to S1
Menu
Parameter Default
Range
values
min.
max.
PID-control S1
Setpoint
250.0 µS/cm
- inf
+ inf
PID-control S1
Range
50.0 µS/cm
- inf
+ inf
PID-control S1
PID-control S1
PID-control S1
Manual Reset
I-time
D-time
0%
3600 s
0 s
0%
1 s
0 s
100%
3600 s
60 s
Duty cycle
Pulse freq.
mA1 (simulate)
DC period time
Max. pulse freq.
Expire time
10 s
70 p/min
0.0 s
1 s
1 p/m
0 s
1800 s
120 p/m
1800 s
Alarm S1
Alarm S2
Alarm S1
Setpoint
Setpoint
Hysteresis
450.0 µS/cm (high)
50.0 µS/cm (low)
4.500 µS/cm
- inf
- inf
0 mS/cm
+ inf
+ inf
+ inf
Alarm S1
Alarm S1
Delay Time
Expire Time
0.2 s
0.0. s
0 s
0 s
+ inf
1800 s
Hold
Hold
Fixed value mA1
Fixed value mA2
12 mA
12 mA
3.6 mA
3.6 mA
21 mA
21 mA
IM 12D08N05-01E
28
5-8. Fail
A fail contact is energized when a fail situation
occurs. Fail situations are configured in secton
5-12. For SOFT Fails the contact and the display on LCD are pulsating. For HARD Fails the
contact and the display on LCD are energized
continuously. Only contact S4 is programmed
as a fail-safe contact. This means that contact
S4 will be de-energized when a fail situation
occurs.
Hard Fail Only
The contact reacts to Hard Fails Only
Hard + Soft fail
The contact reacts to Hard and Soft Fails
5-9. Simulate
The contact can be switched on/off or a
percentage of output can be simulated. On/Off
enables the user to manually switch a contact
on or off. The percentage is an analogue value
and represents the on time per period. The
Duty Cyde Period time (see figure 5-4) is used
as a period for percentage simulation. Note that
the (simulated) settings of the contacts become
visible in measuring mode and after HOLD has
ended c.q. has been overruled. A warning is
activated in case of a simulated output contact.
5-10. Water for Injection Monitoring
(USP 645 and EU 0169).
Setting up EXA SC450 for WFI monitoring
1. A function “USP limit exceeded” is defined
as an error code on sec. 5-12, Errors 2/3.
This can be set to off/warn/fail according
to your requirement. This function can
be modified by the function “USP safety
margin” in %. This is a percentage
of the WFI conductivity value at that
temperature that serves as safety margin.
This is independent of what is being
measured. The display shows this error
when the water quality exceeds the WFI
conductivity limits as set in stage 1.
2. We have introduced uncompensated
conductivity in the DISPLAY menu. In the LCD
display the user can read the temperature
and the raw conductivity to compare his water
quality with the WFI table.
3. We have added a USP function to the
contact allocation. The contact output S1
can be selected as USP alarm if the function
“USP limit exceeded” has been selected.
Commissioning
IM 12D08N05-01E
The contact closes when the USP limit is
reached.
3.5
μS/cm
3.0
USP
Safety
Margin
2.5
2.0
1.5
1.0
0.5
0.0
25
50
75
100
˚C
Figure 5-6. USP Safety Margin
Limit of uncompensated conductivity as function
of temperature as defined for WFI. USP safety
margin set as 20 % will close the contact at 80 %
of the conductivity value at all temperatures.
For example, if the temperature is 64 ºC. and
the safety margin is adjusted for 20%, then the
contact closes at 0.8 x 2.2 μS/cm. = 1.76 μS/cm.
(2.2 μS/cm is the WFI limit at 64ºC). In resistivity
mode the contact will close at an uncompensated
resistivity of 1/1.76 μS/cm. = 0.568 Mohm.
Recommended Commissioning settings when
monitoring WFI in a > 80 ºC WFI installation.
Measurement Set up
Measure
Conductivity only
Temp Compensation
automatic
Conductivity 1
None
Error Configuration (Errors 2/3)
USP limit exceeded
Warn
Output Setup
S1
USP safety margin
S2
Parameter
Setpoint
Direction
Delay Time
Expiry Time
USP
10 %
Alarm
Temperature
80 C
Low
0.2 s
0 (disabled)
5-11. Input contacts
The terminal of the SC450G provides for an input contact (see Figure 3-7). This input contact
can be used to switch the range of the outputs.
The range can be increased by 1 decade.
This is available for only mA1 output.
29
S2, S3, S4 Similar structure to S1
5
MENU STRUCTURE COMMISSIONING
Disabled
Factor 10 mA1
Menu
Parameter Default
Range
values
min.
max.
Simulation
100%
Percentage
50%
0%
IM 12D08N05-01E
30
5-12. Error configuration
Errors 1/3 ~ 3/3
Errors are intended to notify the user of any
unwanted situations. The user can determine
which situations should be classified as:
FAIL, immediate action is required. The process variable is not reliable.
WARN, the process variable processes by the
converter is still reliable at this moment, but
maintenance is required in the near future.
“FAIL” gives a flashing “FAIL” flag in the main
display. The contact configured as FAIL
(Commissioning >> output setup)
will be energized continuously. All the other
contacts are inhibited. A Fail signal is also
transmitted on the mA-outputs when enabled
(burn high/low).
(Commissioning >> output setup)
“WARN” gives a flashing “WARN” flag in the
display. The contact configured as FAIL is
pulsed. All the other contacts are still functional,
and the converter continues to work normally.
A good example is a time-out warning that the
regular maintenance is due. The user is notified, but it should not be used to shut down the
whole measurement.
5-13. Logbook configuration
General
Logbook is available to keep an electronic
record of events such as error messages, calibrations and programmed data changes.
By reference to this log, users can for instance
easily determine maintenance or replacement
schedules.
In “Configure Logbook” the user can select
each item he is interested in to be logged
when the event occurs. This can be done for
three separate logbooks. Each logbook can be
erased individually or all at once. Enable the
”Warn if Logbook full” when you would like to
be warned when the logbook is almost full.
The content of the logbook(s) can also be
retrieved from the converter using the “EXAxt
Configurator” software package which can
be downloaded from the Yokogawa Europe
website.
IM 12D08N05-01E
Flashing “Fail” flag in main display
Flashing “Warn” flag in main display
31
Errors 2/3
Temperature too high
Temperature too low
Polarization detect
Calibr. time exceeded
USP limit exceeded
Warn
Warn
Warn
Off
Off
Enter
5
MENU STRUCTURE COMMISSIONING
Errors 2/3
Temperature too high
Temperature too low
Polarization detect
Calibr. time exceeded
USP limit exceeded
Warn
Warn
Warn
Off
Off
Enter
Menu
Parameter Default
Range
values
Low
High
Errors1/3
Errors1/3
Errors1/3
Errors1/3
2.000 S
2.000 S
10MΩ
10MΩ
Cond. High Limit
Cond. Low Limit
Res. Low Limit
Res. Low Limit
250 mS
1.000 µS
4Ω
1MΩ
>0 mS
>0.00 µS
>0
>0
IM 12D08N05-01E
32
5-14. Advanced setup
Defaults
The functionality of the EXAxt allows to save
and load defaults to come to a known instrument setting. The EXAxt has both factory and
user defined defaults.
After a “load default” the instrument will reset.
The following parameters are not included in
the defaults
1. X-axis timing
2. Auto return (10 min / disabled)
3. Tag
4. Passwords
5. Date and time
6. Language
7. The contents of all logbooks
8. HART parameters
(address, tag, descriptor, message)
Tag
A tag provides a symbolic reference to the
instrument and is defined to be unique
throughout the control system at one plant site.
A tag can contain up to 12 characters. If the
instrument is purchased with the /SCT option,
the TAG is pre-programmed with the specified
tagnumber.
Passwords
Calibration and Commissioning may be separately protected by a password. By default both
passwords are empty. Entering an empty password results in disabling the password check.
A password can contain up to 8 characters.
When a password is entered for the calibration
and commissioning a 4-digit operator ID can be
entered. One can also leave the ID empty.
Date/time
The Logbooks and trend graph use the clock/
calendar as reference. The current date and
time is set here. The current time is displayed
in the third “zoom” menu.
Note! The fixed format is YYYY/MM/DD HH:
MM:SS
HART
The address of the EXAxt in a HART network
can be set. Valid addresses are 0...15.
IM 12D08N05-01E
Factory adjustment
This menu is for service engineers only.
This section is protected by a password.
Attempting to change data in the factory adjustment menu without the proper instructions
and equipment, can result in corruption of the
instrument setup, and will impair the performance of the unit.
33
5
MENU STRUCTURE COMMISSIONING
Menu
Parameter Default
Range
values
Low
High
HART
15
Network address
0
0
IM 12D08N05-01E
34
5-15. Display setup
Main display
The main display consists of three lines with
Process Values. Each line is user definable
with the restriction that each line should have
a different Process Value. The default settings
can be defined here. By pressing one of the
two smaller process values, this will become
the main process value in the main screen.
Autoreturn will cause the main display to go to
default setting.
See also 4-6 Secondary to Primary Value
display Switch.
Note! Configuration possibilities in the
main and secondary display lines are
determined by the choices made in
the menu measurement
Measurement setup >> Measurement
Additional text
Each process value can be given an additional
text containing up to 12 characters per text.
This text is displayed on the main display next
to the process value. This way the user can
distinguish separate measurements.
X-axis timing
The time range of the trend graph can be set
from 15 minutes up to 14 days.
Y-axis limits
The ranges for each measurement need to be
set according the application.
Auto return
When Auto return is enabled, the converter
reverts to the measuring mode (main display)
from anywhere in the configuration menus,
when no button is pressed during the set time
interval of 10 minutes. The HOLD flag will be
cleared and all outputs will function normally!
IM 12D08N05-01E
35
5
MENU STRUCTURE COMMISSIONING
Menu
Parameter Default
Range
values
Low
High
Y-axis
Y-axis
Y-axis
Y-axis
Y-axis
Y-axis
+
+
+
+
+
+
Conduct low
Conduct high
Conduct 2 low
Conduct 2 high
Temp. low
Temp. high
0 µS/cm
500 µS/cm
0 µS/cm
500 µS/cm
0ºC, 0ºF
100ºC, 100ºF
-
inf
inf
inf
inf
inf
inf
inf
inf
inf
inf
inf
inf
IM 12D08N05-01E
36
6. CALIBRATION
6-1. General
The nominal cell constant of a conductivity sensor is determined at the construction
stage, because it is a factor set by the size
of the electrodes, and their distance apart. A
conductivity sensor does not change its cell
constant during operation, as long as it remains
undamaged, and clean. It is therefore vital that
in any calibration check the first step should be
to clean the sensor, or at least check its cleanliness. After cleaning ensure that the sensor is
carefully rinsed in distilled water to remove all
traces of the cleaning medium.
In the commissioning menu, the original sensor
configuration will include the programming
of the cell constant defined for the sensor at
manufacture. Follow the routing below to the
setup screen :
Commissioning >> Measurement setup >>
Configure sensor
The Calibration menu of the SC450G is
provided for fine tuning the sensor setup, and
checking and verification after a time in service.
Where 1st and 2nd compensations are referred
to in this part of the menu, these provide alternatives for the “wet” calibration, designed to
give the user the greatest flexibility.
This does not mean that two cell constants can
or should be calibrated, they are alternative
routes to the same end!
6-2. Cell constant manual
The intention of this calibration routine is to fine
tune a sensor for which only the nominal cell
constant is known, or recalibrate a sensor that
has been changed (or damaged) in the course
of operation. Choose 1st or 2nd compensation
to suit the calibration solution used. The solution should be prepared or purchased, meeting
the highest standards of precision available.
Allow the sensor to reach stable readings for
both temperature and conductivity before adjusting to correspond to the calibration solution
value. The setting of a cell constant for a new
(replacement) sensor is also possible in this
IM 12D08N05-01E
routine. This avoids the need for entry into the
commissioning mode, which may have another
authorization (password) level.
Regarding how to enter cell constant of the
SC21¨G, SC8SG, and SC4AJ, see “Appendix 5.”
6-3. Cell constant automatic
This routine is built around the test method
described in OIML (Organisation Internationale
de Metrologie Legale). International Recommendation No. 56. It allows the direct use of
the solutions prescribed in the test method, automatically selecting the appropriate temperature compensation. The look up table is used
to find the appropriate conductivity reading for
the measured temperature. See appendix 2 for
OIML solutions
6-4. Air (zero) calibration
With the clean dry cell in open air, the reading
should be zero. The Air cal compensates for
excess cable capacitance, and gives a better
accuracy at low readings. This should be done
for all installations during commissioning. After
some time in service a dirty sensor may well
show a high zero offset because of fouling.
Clean the sensor and try again.
6-5. Sample calibration
With the sensor in situ, a sample can be taken
for laboratory analysis. Sample calibration
records the time and reading, and holds these
in memory until the analysis has been completed. The laboratory data can then be entered
regardless of the current process value, without
the need for calculations.
6-6. Temperature coefficient calibration
Simply input the solution conductivity at reference temperature (TR), after the sensor is
allowed to stabilize at elevated temperature.
EXAxt SC450G will calculate the temperature
coefficient for you. The ideal temperature for
this calibration, is the normal process value
(TP). For good calibrations, the minimum span
(TP TR) should be at least 2ºC.
Note that the Temperature Compensation
should be set to TC first.
37
6-7. Temperature calibration
In order to make the most accurate measurements, it is important to have a precise temperature measurement. This affects the display of
temperature, and the output signal when used.
More important, however, is the temperature
compensation, and calibration accuracy.
The temperature of the sensor system should
be measured independently with a high precision thermometer. The display should then be
adjusted to agree with the reading (zero offset
calibration only). For best accuracy this should
be done as near to the normal operating temperature as possible.
6-8. Operation of hold function during calibration
EXAxt SC450G has a HOLD function that will
suspend the operation of the control/alarm
relays and mA-outputs.
c) Solution calibration of SC demands
laboratory technical skills
d) Solutions can be used to give a fair
calibration check at higher conductivity
e) S
olutions can NOT be used to check
calibration at low conductivity.
f)
Low conductivity solutions <10µS/cm
absorb CO2 from the air very fast
g) Low conductivity measurement must be
made only with air excluded
h) Apparatus must be scrupulously clean to
avoid contamination
i)
Sensor linearity is never a problem for
lower values
j)
A dirty sensor is prone to polarization
k) Polarization shows as a low side error at
higher conductivity
l)
A dirty sensor will often read perfectly at
low conductivity
m) Wet calibration tests are best done
towards the top of a sensor’s range
n) If the system responds correctly to the
highest trip point, all is well
IM 12D08N05-01E
CALIBRATION
Press HOLD button on mainscreen, to remove
the HOLD.
The route for HOLD setup is
Commissioning >> Output setup>> Configure Hold
b) There are no good calibration solutions
(like p
H buffer solutions)
6
During calibration, the user may choose to
enable HOLD so that the output signals are
frozen to a “last” or “fixed” value. Some users
will choose to leave the outputs “live” to record
the calibration event. This has implications
for pharmaceutical manufacture, for example,
where an independent record of calibrations is
mandatory.
6-9. General comments on SC calibration
a) SC sensors experience no drift except if
they are damaged or dirty
38
7. MAINTENANCE
7-1. Periodic maintenance
The converter requires very little periodic maintenance, except to make sure the front window
is kept clean in order to permit a clear view of
the display and allow proper operation of the
touchscreen. If the window becomes soiled,
clean it using a soft damp cloth or soft tissue.
To deal with more stubborn stains, a neutral
detergent may be used.
When you must open the front cover and/or
glands, make sure that the seals are clean and
correctly fitted when the unit is re-assembled
in order to maintain the housing’s weatherproof
integrity against water and water vapor.
Note! Never use harsh chemicals or solvents.
In the event that the window does
become heavily stained or scratched,
refer to the parts list (Chapter 10) for
replacement part numbers.
Battery
The EXAxt converter contains a logbook
feature that uses a clock to provide the timings.
The instrument contains a lithium cell (battery)
to support the clock function when the power
is switched off. The cell has an expected working life of 10 years. Should this cell need to
be replaced, contact your nearest Yokogawa
service center.
Fuse
There is a circuit board mounted fuse protecting the instrument. If you suspect that this
needs to be replaced, contact your nearest
Yokogawa service center.
7-2. Periodic maintenance of the sensor
Note! Maintenance advice listed here is
intentionally general in nature. Sensor
maintenance is highly application
specific.
In general conductivity/resistivity measurements do not need much periodic maintenance.
If the EXAxt indicates an error in the measurement or in the calibration, some action may be
needed (ref. chapter 8 troubleshooting).
When a 2-electrode sensor has become fouled
an insulating layer may be formed on the
surface of the electrodes and consequently, an
apparent increase in cell constant may occur,
giving a measuring error.
IM 12D08N05-01E
This error is:
2 x Rv/Rcel x 100 %
where:
Rv = the resistance of the fouling layer
Rcel = the cell resistance
Note! Resistance due to fouling or to
polarization does not effect the
accuracy and operation of a 4electrode conductivity measuring
system. If an apparent increase in cell
constant occurs cleaning the cell will
restore accurate measurement.
7-3. Cleaning methods
1. For normal applications hot water with
domestic washing-up liquid added will be
effective.
2. For lime, hydroxides, etc., a 5 ...10%
solution of hydrochloric acid is
recommended.
3. Organic contaminants (oils, fats, etc.) can be
easily removed with acetone.
4. For algae, bacteria or moulds, use a solution
of domestic bleach (hypochlorite).
* Never use hydrochloric acid and bleaching
liquid simultaneously. The release of the very
poisonous chlorine gas will result.
7-4. LCD adjustment
Contrast adjustment
During the life of the analyzer the contrast of
the display may fade. The contrast can be adjusted using the potentiometer on the backside
of the LCD board. This adjustment must be
done only by Yokogawa’s service personnel.
The position is shown on this picture. For units
manufactured after July 2007, the potentiometer is placed behind the little hole in the LCD
bracket as shown in Figure 3-4 on page 7.
For units manufactured between April 2006
and April 2007, the potentiometer is located as
shown right.
39
Touchscreen adjustment
CAUTION
A few years after using, the touchscreen
may deviate from the correct position due
to aging deterioration of the touchscreen.
When that happens, turn off power then on
again. The touchscreen will be calibrated
automatically to the correct touch position
at power on. It is recommended to turn
off power then on again when periodic
maintenance.
CAUTION
Do not turn on power with the touchscreen
pressed, otherwise inaccurate touch position
will occur. If it occurs, leave the touchscreen
unpressed, turn off power then on again.
The touch position will be accurate.
7 MAINTENANCE
IM 12D08N05-01E
40
8. TROUBLESHOOTING
8-1. General
The EXAxt is a microprocessor-based analyzer
that performs continuous self-diagnostics to
verify that it is working correctly. Error messages resulting from faults in the micro-processor systems itself are monitored. Incorrect
programming by the user will also result in an
error, explained in a message, so that the fault
can be corrected according to the limits set in
the operating structure. The EXAxt also checks
the sensor system to establish whether it is still
functioning properly. In the main display screen
is a “Status Information” button that will show
8-3. Polarization check
The EXAxt SC450G performs on-line monitoring to detect polarization. This is an early
indicator for sensor fouling. The detection
of polarization in the measurement gives a
warning of the onset of sensor coating, before
significant measuring errors build up.
For information
8-5. Prediction of cleaning needed
The date when the next maintenance is needed
is calculated from on-line polarization checks.
The trend of polarization measurements on the
sensor is used to calculate when to tell the user
when to clean the sensor.
For warning - a potential problem is diagnosed, and the system should be checked.
For FAIL, when the diagnostics have confirmed
a problem, and the system must be checked.
This button gives access to a status report
page, where “The most applicable error” will
be displayed. (“No errors” is displayed during
proper operation)
Explanation >> Description or error message
and possible remedies
Advanced troubleshooting >> Error code
screen that is used in conjunction with the
service manual. This data will also be needed
in the event that you request assistance from a
Yokogawa service department.
What follows is a brief outline of the EXAxt
troubleshooting procedures including possible
causes and remedies.
8-2. Calibration check
The EXAxt SC450G converter incorporates a
diagnostic check of the adjusted cell constant
value during calibration. If the adjusted value
stays within 80-120 % of the factory value, it is
accepted, otherwise, the unit generates an error message, and the calibration is rejected.
IM 12D08N05-01E
8-4. Predictive maintenance
EXAxt has a unique prediction feature. Calibration, and polarization check data are stored in
software data logbooks. This data is then used to
calculate a prediction for maintenance purposes.
8-6. Poor calibration technique
When the calibration data is not consistent
this fact is used as a diagnostic tool. The
significance of this error message is to require
the user to improve his calibration technique.
Typical causes for this error are attempting
to calibrate dirty sensors, calibration solution
contamination and poor operator technique.
8-7. Error displays and actions
All errors are shown in the “Main Display”
screen, however, the EXAxt makes a distinction between diagnostic findings. The error
messages may be set to OFF, WARN or FAIL.
For process conditions where a particular diagnostic may not be appropriate, the setting OFF
is used. FAIL gives a display indication only of
that the system has a problem and inhibits the
relay control action, and can be set to trigger
the “Burn” function. “Burn-up” or “Burn-down”
drives the mA output signal to 21 mA or 3.6 mA
respectively.
41
9. QUALITY INSPECTION
Quality
Inspection
Standards
1.
SC450G
Conductivity or Resistivity Converter
Scope
This inspection standard applies to the SC450G Conductivity or Resistivity Converter.
2.
Inspection Items
2.1
*2.2
*2.3
2.4
2.5
Insulation resistance test
Dielectric strength test
Sensor signal input test
Temperature indication check
Current output test
Note: Items marked with an asterisk (*) may only be confirmed by a test certificate.
3.
Inspection Methods, Standards and Conditions
z
3.1
Connect the testing circuit as shown in Figure 1. Allow the instrument to warm up for
at least 5 minutes before conducting the tests. For the connections for the insulation
resistance and dielectric strength tests, follow the instructions in Sections 3.1 and 3.2.
Insulation Resistance Test
Apply 500 V DC between the terminals shown below. The insulation resistance must be
100 MΩ or greater.
(1) Between the power supply terminals shorted together (1 and 2) and the protective
earth terminal ( )
(2) Between the contact output terminals shorted together (32, 33, 42, 43, 52, 53, 72 and
73) and the protective earth terminal ( )
(3) Between the current output terminals shorted (62) and the protective earth terminal
( )
Dielectric Strength Test
(1) Apply 1390 V AC, an AC voltage of substantially sinusoidal waveform with a frequency
of 50 Hz or 60 Hz, between the power supply terminals shorted together (1 and 2) and
the protective earth terminal ( ), for at least 2 seconds. The insulation must withstand
this voltage. (The sensed current should be 10 mA.)
(2) Apply 1390 V AC, an AC voltage of substantially sinusoidal waveform with a frequency
of 50 Hz or 60 Hz, between the contact output terminals shorted together (32, 33, 42,
43, 52, 53, 72 and 73) and the protective earth terminal ( ), for at least 2 seconds.
The insulation must withstand this voltage. (The sensed current should be 10 mA.)
(3) Apply 500 V AC, an AC voltage of substantially sinusoidal waveform with a frequency
of 50 Hz or 60 Hz, between the current output terminals shorted (62) and the
protective earth terminal ( ), for at least 2 seconds. The insulation must withstand
this voltage. (The sensed current should be 10 mA.)
3.3
Sensor Signal Input Test
Connect the testing circuit as shown in Figure 1 and set the equipment as follows:
Decade resistance box 1 (temperature simulation input):
1097.3 [Ω]
Decade resistance box 2 (conductivity simulation input):
150 [Ω]
The power supply voltage should be set in accordance with the specifications of the
converter.
QIS 12D08N05-01E
1st Edition: Jul. 2007
IM 12D08N05-01E
9 QUALITY INSPECTION
3.2
42
2/3
This test is done on the “HIF” display of “Factory Mode”.
a. Touch the [Setup] icon.
b. Touch the [Commissioning].
c. Touch the [Advanced setup].
d. Touch the [Factory adjustment].
e. Enter a password.
f. Select the [Factory Mode] in “Key.”
g. Select the [HIF] in “Execute.”
When the resistance of the decade resistance box 2 to the corresponding value
“RANGE1” to “RANGE5” in Table 1 is set, check the data display and the value must be
within the range shown in Table 1.
Table 1
RANGE
RANGE1
RANGE2
RANGE3
RANGE4
RANGE5
Resistance (Ω) of
Decade Resistance Box 2
20
200
2k
20k
200k
Data Display (Ω)
20
200
2k
20k
200k
±0.1
±1
±0.01k
±0.1k
±1k
After the above test is completed, touch the [Exit] to return to the “HIF” display.
3.4
Temperature Indication Check
Following Section 3.3, select the [PT1000] in “Temperature” of the “HIF” display.
In this state, change the resistance of the decade resistance box 1 and check the data
display. The value on the data display must be within the range shown in Table 2.
Table 2
Temperature (°C)
–10
25
240
Resistance (Ω) of
Decade Resistance Box 1
960.9
1097.3
1904.6
Data Display (°C)
–10 ±0.3
25 ±0.3
240 ±0.3
After the above test is completed, touch the [Exit] to return to the “HIF” display.
3.5
Current Output Test
Following Section 3.4, select the [Check] in “mA outputs” of the “HIF” display. “Set value
4.000 mA” appears at the bottom of the display. Select “Next value” in the “Command”
and touch “Enter,” the value on the data display increases in steps of 4 mA. Check the
current outputs 1 and 2 corresponding to the data display, the current output must be
within the range shown in Table 3.
Table 3
Data Display
4
12
20
Current Output (mA DC)
4 ±0.02
12 ±0.02
20 ±0.02
QIS 12D08N05-01E
IM 12D08N05-01E
43
3/3
After all tests are completed,
a. Touch the [Exit] twice to return to the “Service” display.
b. Select “Normal” in “Key”.
c. Touch the [Home] icon to return to the initial display.
POWER
SUPPLY
1
(+ )
2
(- )
SC450G
66
mA2
+
65
62
mA1
TEMP
+
61
300Ω
-
DC +
Ammeter
POWER
SUPPLY
11
12
ELECTRODE
14
13
15
16
300Ω
- DC +
Ammeter
Decade
Resistance
Box 1
Decade Resistance
Box 2
Figure 1 Testing Circuit and Test Equipment
9 QUALITY INSPECTION
QIS 12D08N05-01E
IM 12D08N05-01E
44
IM 12D08N05-01E
45
10. SPARE PARTS
See Customer Maintenance Parts List.
9 SPARE PARTS
IM 12D08N05-01E
46
APPENDICES
Appendix 1
Temperature compensation
The conductivity of a solution is very dependent on temperature. Typically for every 1°C change in
temperature the solution conductivity will change by approximately 2%. The effect of temperature
varies from one solution to another and is determined by several factors like solution composition, concentration and temperature range. A coefficient (α) is introduced to express the amount of
temperature influence in % change in conductivity/°C. In almost all applications this temperature
influence must be compensated before the conductivity reading can be interpreted as an accurate
measure of concentration or purity.
NaCl or standard temperature compensation
From the factory the EXAxt is set with the default of a general temperature compensation function
based on a Sodium Chloride (table salt) solution. This is suitable for many applications and is compatible with the NaCl compensation functions of typical laboratory or portable instruments.
SC temperature compensation.
Table 11-1. NaCl-compensation according to IEC 60746-3 with Tref = 25 °C
Configure calculated temperature coefficient (TC).
Follow routing
Commissioning >> Measurement setup >>
Temp.compensation >> T.C.
Enter the temperature coefficient calculated from the
following formula:
A. Calculation of temperature coefficient factor
( With known conductivity at reference temperature).
α=
α
T
Kt
Tref
Kref
Kt - Kref
T - Tref
X
100
Kref
= Temperature compensation factor in %/°C
= Measured temperature in °C
= Conductivity at T
= Reference temperature
= Conductivity at Tref
IM 12D08N05-01E
47
B. Calculation of temperature coefficient factor
(with two known conductivity values at different temperatures)
Measure the conductivity of the liquid at two temperatures, one below the reference and above
the reference temperature with the temperature coefficient set to 0,00%/°C and use the following
equation to calculate a temperature coefficient (α).
K ref =
K ref =
KT
1+α ( T - Tref)
K1
1+α ( T1 - Tref)
=
K2
1+α ( T2 - Tref)
K1 (1 + α ( T2 - Tref)) = K2 (1 + α ( T1 - Tref))
K1 · α ( T2 - Tref) - K2 · α ( T1 - Tref) = K2 - K1
α=
K2 - K1
K1( T2 - Tref) - K 2 ( T1 - Tref)
Where T1 , T2 : liquid temperature (°C)
K1 : conductivity at T 1 (°C)
K2 : conductivity at T 2 (°C)
Figure 11-1. Conductivity
Calculation example
Calculate the temperature coefficient of a liquid from the following data.
Conductivity 124.5 µS/cm at a liquid temperature of 18.0 °C and a
conductivity 147.6 µS/cm at a liquid temperature of 31.0 °C.
Substituting the data in the above formula gives the following result.
147.6 - 124.5
x 100= 1.298 %/C
124.5(31.0 - 25) - 147.6(18.0 - 25)
Set the temperature coefficient in the SC450G converter.
α=
A APPENDICES
IM 12D08N05-01E
48
Checking
When the temperature coefficient already set is accurate, the conductivity to be displayed must be
constant regardless of liquid temperature. The following check will make sure that the temperature
coefficient already set is accurate.
If, when the liquid temperature is lowered, a larger conductivity value is indicated, the temperature
coefficient already set is too small.
The opposite also applies. If a smaller conductivity value is indicated, the temperature coefficient
already set is too large. In either case, change the temperature coefficient so that the conductivity
no longer changes.
Matrix compensation
The compensation matrix is a table of temperature and conductivity values at differing concentrations. These values are used to calculate the temperature compensation applicable for a particular
solution. Choose the component that you will be measuring in your application, and where appropriate the concentration range. The EXAxt will do the rest.
By following the routing:
Commissioning>> Measurement setup>> Temp.compensation>> Matrix
you gain access to the Matrix selection area.
Matrices are available for the common mineral acids and bases. In addition Ammonia and Morpholine are included. In short by using the matrix method, specialist compensation is available for the
majority of applications in the power industry, water treatment, and chemical manufacturing. The
following matrices are available initially, but as with all Yokogawa products, we are continually striving to improve both the quality and technological content. Further solutions will be added to this list.
Ammonia
0..50 ppb
0..90˚C
Ammonia
15..30%
10..50˚C
Morpholine
0..500 ppb
0..90˚C
0..27%
0..100˚C
H2SO4
39..85%
-18..116˚C
H2SO4
93..100%
10..90˚C
NaOH
0..15%
0..100˚C
H2SO4
NaOH
25..50%
0..80˚C
0..200 ppb
0..100˚C
HCI
0..18%
-10..65˚C
HCI
24..44%
-20..65˚C
HCl
HNO3
0..25%
0..80˚C
HNO3
35..80%
-16-60˚C
IM 12D08N05-01E
49
Temperature compensation matrix.
1. A minimum number of values is required to make interpolation possible.
The highlighted values marked as
are mandatory to enter.
Sol1
....
Solx
....
Sol10
Concent
C1
Tref
S1Tr
T1
S1T1
....
Tx
....
T10
S1T10
SxTx
C10
S10Tr
S10T1
S10T10
2. Tref (reference temperature) is defined in the Temperature Compensation menu. If Tref is
between T1 and T10 then the value of Tref needs to be entered as Tx (T2....T9)
Sol1
....
Solx
....
Sol10
3.
Concent
C1
Tref
S1Tr
T1
S1T1
C10
S10Tr
S10T1
....
Tx
.... T10
S1T10
S10T10
For every SxTx that is entered the following values become mandatory to enter:
Cx, SxTr, SxT1, SxT10 and Tx
Sol1
....
Solx
....
Sol10
Concent
C1
Tref
S1Tr
T1
S1T1
Cx
SxTr
SxT1
C10
S10Tr
S10T1
....
Tx
.... T10
S1T10
SxTx
SxT10
S10T10
The matrix can be cleared before entering new
values. Next new matrix values can be entered
as described above. The EXAxt can interpolate
the matrix. During this process it will check if
the matrix is completely ascending/descending. This is necessary as otherwise the lookup
function can give two results for one temperature. If an error is found, the EXAxt will specify
the location of the error as shown in the user
interface screen “user defined 1/2”.
A APPENDICES
The backspace key should be used for deleting an individual matrix value.
An empty value is shown as
IM 12D08N05-01E
50
TDS readings
The concept of Total Dissolved Solids (TDS) is
widely used as an indication of the total solute
concentration in water and is a widely accepted water quality standard. The determination
is simply made by evaporation and weighing
the residue. As the heat treatment converts
the chemical properties of the solution, some
solutes that contribute to the total conductivity will not be found in the residue. Also some
combinations of solutes will become volatile
at the drying temperature. The same applies
for those solutes that do not contribute to the
conductivity but do end up in the residue.
When one (or a few) solute is dominant in the
solution, these problems may be neglected
and TDS will have a direct correlation with the
conductivity. Next is a graph that shows the
correlation between the specific conductance
and concentrations (by weight) for six salts.
The factor by which the conductivity should be
multiplied to come to a TDS reading ranges
from 0.4 to 0.7 up to a conductivity of 500 µS/
cm. At a conductance of approximately 3000
µS/cm the range would be from 0.5 and 1.0.
The EXAxt 450 measures conductivity. This
value can be used to display TDS values on
the main display. One should use the additional concentration table. Please refer to section
5-5 for concentration measurement. The unit
for the TDS reading can be set to ppm or ppb.
IM 12D08N05-01E
Specific conductance, µS/cm, 25°C
Appendix 2
7000
KCI
NaCI
CaCI2
6000
Na2SO4
MgCI2
5000
4000
MgSO4
3000
2000
1000
0
1000
2000
3000
4000
5000
Concentration, mg/L
51
Appendix 3
Calibration solutions for conductivity
Note!
This section should be read in conjunction with the calibration section (Chapter 6) and the maintenance section (Chapter 7). The calibration (cell constant) of a sensor does not change unless the
sensor is damaged.
It can also appear to change because of coating of the electrodes, or partial blockage. Because
these changes should be handled as described in the maintenance section, it does not make
sense to regularly recalibrate the SC450G.
A calibration check, however, is another matter. When the objective is clearly defined as a diagnostic exercise a regular check can bring an extra level of security and confidence to the measurement.
Sensor damage, and/or coatings can be difficult to see and the calibration check can confirm their
presence, by a deviation from the known solution conductivity. The remedial action should be to
clean the sensor, and carefully check for blockage or damage (not simply to recalibrate).
Higher conductivity solutions should be used where possible. The lower the conductivity of the test
solution, the easier it is to contaminate. Carbon dioxide from the air can be quickly absorbed to
cause an error. All containers must be suitably clean, and all materials suitably pure. Outside of a
well-equipped laboratory these conditions are hard to meet.
Also note that the check must be performed with due regard to the cell constant of the sensor,
as it limits the effective working range. The documentation provided with the sensor should be
consulted to determine its maximum working value. By using a checking solution close to the
upper range limit of the sensor, the detection of contamination is better. A slightly contaminated
sensor may read perfectly at lower conductivity, but show significant errors at higher values. This
is caused by the early onset of polarization, the factor that limits the upper range of the sensor
in any case. Polarization is seen in contaminated sensors, and it is for this reason that there is a
sophisticated polarization check built into the SC450G. Details of this diagnostic tool are found in
the troubleshooting section (Chapter 8).
EXAxt SC450G is programmed with the following table of conductivity of Potassium Chloride (KCl)
solutions at 25°C. This is used in the Automatic Cell Constant setting calibration feature. (See
chapter 6 on calibration) The table is derived from the Standards laid down in “International Recommendation No. 56 of the Organisation Internationale de Métrologie Legale”.
Table 11-2. KCl values at 25 °C
mol/l
A APPENDICES
IM 12D08N05-01E
52
If it is more convenient, the user may make solutions from Sodium Chloride (NaCl or common
table salt) with the help of the following relationship table. This table is derived from the IEC norm
60746-3.
Table 11-3. NaCl values at 25 °C
Weight %
0.001
0.003
0.005
0.01
0.03
0.05
0.1
0.3
0.5
1
3
5
10
mg/kg
10
30
50
100
300
500
1000
3000
5000
10000
30000
50000
100000
Conductivity
21.4 µS/cm
64.0 µS/cm
106 µS/cm
210 µS/cm
617 µS/cm
1.03 mS/cm
1.99 mS/cm
5.69 mS/cm
9.48 mS/cm
17.6 mS/cm
48.6 mS/cm
81.0 mS/cm
140 mS/cm
Note!
For resistivity measurement the standard resistivity units of the calibration solution can be calculated as follows:
R = 1000/G (kΩ•cm if G = µS/cm)
Example: 0.001% weight
IM 12D08N05-01E
R = 1000/21.4 = 46.7 kΩ•cm
53
Appendix 4
Sensor selection for the EXAxt SC450G
Yokogawa supplies the following sensors, and their application is briefly described. All are compatible with the EXAxt SC450G, and the user must be aware of the sensor configuration for 2 or 4
electrodes, in order to set the converter correctly.
SC42-SP34 (&SX42-SX34) cc = 0.01 cm-1
2-electrode sensor
This sensor is the first choice for pure and ultra pure water measurements. From 0.055µS/cm up to
100µS/cm this is a perfect choice. The measuring range extends to at least 1000µS/cm, but there
are better choices in that range.
Almost exclusively used in flow fittings and sub-assemblies. FF40 and FS40. (The -SX sensors are
screwed or flanged direct into the process pipework or tanks.)
SC4A-S(T)-XX-002...... cc = 0.02 cm-1 2-electrode sensor
This sensor is a good choice for pure and ultra pure water measurements. From 0.055µS/cm up
to 50µS/cm this is a good choice. The measuring range extends to 500µS/cm, but there are better
choices in that range. The “SC4A-” sensors make up a range that use compression adapters or
specialized fittings, including retractable.
SC42-SP24 (&SX42-SX24) cc = 0.1cm-1
2-electrode sensor
This sensor should be used for clean and pure water measurements. From 0.2µS/cm up to 1
mS/cm the sensor is fine, with measurement up to 10mS/cm possible, but there are better choices
in that range. Mounting of these sensors is as for the ....-SP(X)34 series.
SC4A-S(T)-XX-010...... cc = 0.1 cm-1 2-electrode sensor
This sensor is a good choice for clean water measurements. From 0.5µS/cm up to 200µS/cm this
is a good choice. The measuring range extends to 2000µS/cm, but there are better choices in that
range. The “SC4A-” sensors make up a range that uses compression adapters or specialized fittings, including retractable.
SC42-EP15(D)
cc = 1 cm-1
2-electrode sensor
This sensor is intended as a handy, short sensor. It is easy to fit in a branch (tee) on a pipeline,
and is good for measurements where the precision is less critical than a convenient installation.
The D version is particularly suited to food applications because of it’s cleanability. Range of application is from 10µS/cm to 10mS/cm.
SC42-EP14
cc = 1 cm-1
2-electrode sensor
Clean cooling and process water applications are where this sensor is used. The best range of application is 10µS/cm to 10mS/cm. It is better suited to flow through installations in FS40 and FF40
adapters and fittings, but it can also be used in dip tubes - FD40.
IM 12D08N05-01E
A APPENDICES
SC42-EP18
cc = 1 cm-1
4-electrode sensor
Cooling and process water applications are where this sensor is used, including light fouling. The
best range of application is 100µS/cm to 100mS/cm. The four-electrode system copes with some
contamination from the process, and eliminates the tendency for polarization to occur at the higher
conductivity. It is better suited to flow through installations in FS40 and FF40 adapters and fittings,
but it can also be used in dip tubes - FD40.
54
SC42-EP04
cc = 10 cm-1
2-electrode sensor
This sensor is in the program for historical reasons. It makes little sense to select this sensor for
new applications.
It was used mostly for process water and light process solutions. The operating range can be 1to
100 mS/cm with higher readings having a very high risk of polarization. The SC42-EP18 or the
SC42-EP08 (4-el sensors) covers these ranges with much greater certainty of avoiding polarization.
SC42-EP08
cc = 10 cm-1
4-electrode sensor
Process applications are where this sensor is used, including light fouling. The best range of application is 1mS/cm to 500mS/cm. The four-electrode system copes with some contamination from
the process, and eliminates the tendency for polarization to occur at the higher conductivity. The
upper limit is determined more often by chemical compatibility. The epoxy material is attacked
by many of the chemicals that are present in high conductivity solutions. It is better suited to flow
through installations in FS40 and FF40 adapters and fittings, but it can also be used in dip tubes FD40.
SC42-TP08 (FP08)
cc = 10 cm-1
4-electrode sensor
These sensors are for extreme applications. The PTFE or the PVDF used for the body of the sensor gives good chemical resistance for the strongly corrosive applications. These sensors are
intended for use in ranges 10mS to 1000mS/cm. Careful selection of the holder is needed. The
PVDF sub-assembly is an obvious choice for the SC42-FP08, and the PP flow fitting often has sufficient corrosion resistance for the application. In any case selection must be done with due regard
to the process conditions.
The 2-electrode versions (SC42-T(F)P04 do exist, but like the epoxy version (SC42-EP04) they
should be discounted for new applications.
IM 12D08N05-01E
55
Appendix 5, How to enter the cell constant for the SC21¨G, SC8SG, and SC4AJ
When install a new sensor, or replace a new one, put the cell constant manually as below.
(1) In the case that the only cell constant is mentioned on the text plate of the sensor (SC211G,
SC8SG, SC4AJ). Enter the cell constant directly.
(2) In the case that the deviation of a nominal cell constant (± X.X%) is mentioned on the text plate
of the sensor (SC210G).
When the nominal cell constant is 5 cm-1 and the deviation (CORR.% = -1.1) is mentioned:
The cell constant to be entered is calculated as follows:
5 + 5 x (-1.1/100) = 4.945
Enter the cell constant of 4.945 cm-1.
A APPENDICES
IM 12D08N05-01E
56
Appendix 6, HART HHT (275/375) Menu structure
Online menu
Level 1 menu
Process
values
Primary value (SC/RES/Conc.)
Secondary value (Temp.)
Tertiary value (SC/RES/Cond)*
Level 2 menu
Zoom
Zoom sensor
Fact CC
Adj CC
Method SC1
Method SC2*
Pol%
Ohms
USP%
Zoom outputs
mA1 value
mA2 value
S1 perc.
S2 perc.
S3 perc.
S4 perc.
Zoom device
Serial number
(Note)
Software Revision
Device Revision
DD Revision
Logbook
Sensor data
Calibration
Sensor
Pred.Maint
Output data
Settings
mA1
mA2
S1
S2
S3
S4
(Note): A part of the HART device ID (descriptor)
IM 12D08N05-01E
Level 3 menu
Level 4 menu
57
Online menu
Level 1 menu
Most appl. Error
Error description / remedy
Calibration
CC Calibration SC1
CC Calibration SC2
Air Calibration
Sample calibration SC1
Sample calibration SC2
TC Calibration SC1*
TC Calibration SC2*
Temp. Calibration
Hold Instrument
Hold Instrument
Hold Outputs
Hold Off
Commissioning
Measurement setup
Level 2 menu
Meas
Configure sensor
Level 3 menu
Level 4 menu
Sensor type
Meas unit
Fact CC
Measure*
Temp settings
Temp sensor
Temp unit
Temp compensation
Temp comp
Man value*
Ref temp
Method SC1
TC SC1*
Matrix SC1*
Method SC2*
TC SC2*
Matrix SC2*
Calib. settings
Air adjust limit
cc hi limit
cc lo limit
Stab time
cal interval
Concentration
Additional table*
Conc table unit*
A APPENDICES
IM 12D08N05-01E
58
Online menu
Level 1 menu
Level 2 menu
Commissioning
Output setup
mA1 setup
mA2 setup
(similar to mA1)
S1
S2
S3
S4
setup
setup (similar to S1)
setup (similar to S1)
setup (similar to S1)
Level 3 menu
Level 4 menu
Type = control
Func
Process
parameter
PID SP
PID Rng
PID dir
PID MR*
PID I-time*
PID D-time*
Burn
Expiry time
Type = output
Func
Process
parameter
Lin 0%*
Lin 100%*
Burn
Damping time
Type = simulate
Func
Sim. Perc.
Type = Off
Type = control
Func
Process
parameter
Expire time
PID SP
PID Rng
PID dir
PID MR*
PID I-time*
PID D-time*
Analog output
DC period time*
max. pulse freq.*
Type = fail
func
Type = alarm
Func
Process
parameter
alarm SP
alarm dir.
alarm hyst.
alarm delay
expiry time
Type = simulate
func
on/off*
percentage*
Type = hold
func
Type = Off
HOLD setup
HOLD L/F
mA1 fixed *
mA2 fixed *
Hold during cal
Input contact
Configure Input
contact
Error config
Configure error
Off/Warn/Fail
set limits
Logbook config
Sensor logbook
mA logbook
Contact logbook
Erase logbook
Warn logbook full
IM 12D08N05-01E
Calibration
Sensor
Predictive. Maint.
All logbooks
59
Online menu
Level 1 menu
Level 2 menu
Level 3 menu
Level 4 menu
Loop test
Basic setup
Review
Tag
Distributor
Model
Device information
Date
Descriptor
Message
Poll addr
Num resp preams
Model
Distributor
Write protect
Manufacturer
Dev id
Tag
Descriptor
Message
Date
Universal rev
Fld dev rev
Sofware rev
Poll addr
Num req preams
(Note): HART protocol DD files can be downloaded by following URL.
http://www.yokogawa.com/an/download/an-dl-fieldbus-001en.htm
A APPENDICES
IM 12D08N05-01E
60
Appendix 7, Control drawing for FM approval
IM 12D08N05-01E
Customer
Maintenance
Parts List
Model SC450G [Style: S2]
Conductivity / Resistivity
Converter
7, 10
3
9
2
11
5c
5a, 5b
8
6a, 6b
12
1a, 1b
Item
*1a
*1b
*2
*3
Part No.
K9676GA
K9676HA
K9677EA
K9676MA
*5a
*5b
*5c
*6a
*6b
*7
8
9
10
*11
12
K9676MX
K9676MW
K9676MY
A1108EF
A1111EF
K9676BE
K9676BU
K9676DL
K9676BT
K9676CM
—
K9171SU
K9316AF
K9676BC
Qty
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
Description
Power board assembly AC version
Power board assembly DC version
Main board assembly SC version
LCD module
Cable assembly (3 core)
Cable assembly (10 core)
Cable assembly (shield)
Fuse AC version (1 pcs.)
Fuse DC version (1 pcs.)
Cover assembly without, screws and hingepins
Cable glands assembly (6 pcs. M20)
Stainless tagplate blank
Screw assembly to fix cover (M4 screws, washer, O-ring, hingepins)
Housing assembly polyurethan baked finish
Adapter assembly for conduit work
For G1/2 screw when /AFTG specified
For 1/2NPT screw when /ANSI specified
For M20 screw when /AM20 specified
*) Do not exchange these parts. Call service personnel.
©Copyright 2007, 4th Edition: Oct, 2011 (YK)
Subject to change without notice.
CMPL 12D08N05-02E
2
Pipe/Wall Mounting Hardware (Option code: /U)
Panel Mounting Hardware (Option code: /PM)
7
2
8
3
3
9
4
5
6
1
UNIVERSAL MOUNT SET (Option code : /UM) includes
both “/U” and “/PM”.
3
10
Awning Hood (Option code: /H5)
CMPL 12D08N05-02E
Item
1
2
3
4
5
Part No.
K9171SS
Y9608KU
D0117XL-A
K9171SY
K9171SX
Qty
1
4
1
1
1
Description
Mounting Set (/U)
Screw
U-Bolt Assembly
Plate
Bracket
6
7
8
9
K9171ST
Y9520LU
K9171SW
Y9608KU
1
2
2
4
Mounting Set (/PM)
Screw
Bracket
Screw
10
K9676BA
1
Awning hood assembly (/H5)
4th Edition : Oct. 2011 (YK)
Revision Record
Manual Title :
Model SC450G [Style: S2] Conductivity/Resistivity Converter
Manual Number : IM 12D08N05-01E
Edition
1st
2nd
Date
Jul. 2007
Sep. 2007
3rd
Aug. 2008
4th
Mar. 2012
Remark (s)
Newly published
Back-side of cover, note added; p1, FM approval description of
Figure 1-1 changed; p7, some CAUTION of Figure 3-4 and the
touchscreen added; p16, sec. 4-3-13 Serial number changed; p19,
parameter values and some screen corrected; p21, Made some
revisions (parameter values and some screen corrected; p23,
parameter values and some screen corrected; p25, parameter
values corrected; p29, parameter values corrected; p33, made
some revisions; p37, some CAUTION of the touchscreen
added; p39 to p42 Sec. 9 QUALITY INSPECTION inserted; p43
Sec. 10 SPARE PARTS section and page moved; p44 to p57
APPENDICES page moved (p53, note of serial number added;
p57, APPENDIX 6, Control drawing for FM approval added);
CMPL 12D08N05-02E, 1st Edition, made some revisions.
Revisions: Back-side of cover, note illustration added; p3,
Japanese added to display language; p4, option codes /U, /
PM, /H5, /AFTG, /ANSI, /AM20 added to Model and codes; p5
to 6, Layout changed (Figure 3-1. moved and changed, because
external dimensions for awning hood /H5, conduit adapter /AFTG,
/ANSI, /AM20 added); p7, Figure 3-4 title modified; p8, conduit
adapter work added to subsection 3-2-2; p9 to 11, Layout changed
(descriptions after conduit adapter work moved.); p16, Subsection
4-3-13. Serial number --> HART ID. changed; p20 to 35 Layout
changed (illustration of submenu screen placed on appropriate
page); p41, some error corrected; p47, range value corrected;
p49, some error corrected; p50, Table 11-2 modified; p53, C.C.
range modified; p57, Note of HART protocol DD files URL added;
Customer Maintenance Parts List CMPL12D08N05-02E revised
to 2nd edition, because Part No. for option codes /U, /PM, /H5, /
AFTG, /ANSI, /AM20 added.
Revisions: PREFACE, Addition of “How to dispose the batteries”;
p1, Addition of KC mark; p4, Some revision of O) Safety and
EMC conforming standards (description for EMC revised), addition
of KC mark; p9, Caution mark position on Section 3-3-1 changed;
p15, HART communication mark added to Figure 4-1; p28, Some
revision of Section 5-10. Water for Injection Monitoring, (USP 645
and EU 0169). Setting up EXA SC450 for WFI monitoring; p36,
Some revision of Section 6-2 “Cell constant manual”; p38 to 39,
Section 7-4 “Contrast adjustment” changed to “LCD adjustment”,
some caution added; p50, Some revision of Table 11.2 (Weight %
deleted), and change of page; p55, Addition of “Appendix 4 How
to enter the cell constant for the SC21¨G, SC8SG, and SC4AJ.”;
p56 to 60, Appendix 5 changed to Appendix 6, and Appendix
6 changed to Appendix 7; Customer Maintenance Parts List
CMPL 12D08N05-02E, revised to 4th Edition, some of illustration
changed.
User’s
Manual
Model SC450G [Style: S2]
Conductivity / Resistivity
Converter
Supplement
Thank you for selecting "Model SC450G [Style: S2] Conductivity / Resistivity Converter."
The user's manual, IM 12D08N05-01E, 4th Edition, supplied with the product has been amended
as follows. Please replace the corresponding pages in your copy with the attached, revised pages.
Revisions:
- Page i to ii, Addition of Product Disposal and Authorized Representative in EEA.
- Page 1, Changing of Nameplate, and addition of Note.
- Page 2 to 4, S
ome spell revision of standard specification.
Some spell error corrections and addtion of specification description.
Changing of Shipping details ; from Package size to Dimension.
Deleting of CE mark, and addition of note.
- Page 5 to 6, S
ome revision of Figure 3-1 and Figure 3-3.
(addtion of M6-screw position for wall mounting).
- Page 7, Changing of Cables that withstand temperatures.
- Page 10 to 11, Addition of coments of 3-3-3, 3-3-4, 3-4-2, and 3-5-2.
- Page 20, S
ome caution added when measuring unit change in the 5-1. Configure sensor.
And addition of description about a sensor exchange.
- Page 41 to 44, QIS 12D08N05-01E, QIC-12D08N05-01 revised
(some change of insulation & dielectric strength test).
- CMPL 12D08N05-02E, Some revision of P/N (item 2).
Additional information on a product certification is as below;
- An exclusive User’s Manual might be attached to the products whose suffix codes or option
codes contain the code “Z”. Please read it along with this manual.
- RoHS:
EN 50581:2012 (Style 2.34 or newer)
- u CE marking products
n Identification Tag
This manual and the identification tag attached on a packing box are essential parts of
the product.
Keep them together in a safe place for future reference.
n Users
This product is designed to be used by a person with specialized knowledge.
©Copyright 2012, 7th Edition: Apr., 2017 (YK)
Subject to change without notice.
IM 12D08N05-01E
4th Edition
i
PREFACE
Electrostatic discharge
The SC450G converter contains devices that
can be damaged by electrostatic discharge.
When servicing this equipment, please observe
proper procedures to prevent such damage.
Replacement components should be shipped in
conductive packaging. Repair work should be
done at grounded workstations using grounded
soldering irons and wrist straps to avoid
electrostatic discharge.
DANGER
Installation and wiring
The SC450G converter should only be used with
equipment that meets the relevant IEC, American
or Canadian standards. Yokogawa accepts no
responsibility for the misuse of this unit.
CAUTION
The Instrument is packed carefully with shock
absorbing materials, nevertheless, the instrument
may be damaged or broken if subjected to strong
shock, such as if the instrument is dropped.
Handle with care.
WARNING
• Do not use an abrasive or organic solvent in
cleaning the instrument.
• Do not modify the SC450G converter.
• Substitution of components may impair
suitability for Division 2.
Do not remove or replace while circuit is live
unless area is know to be non-hazardous.
Explosion Hazard – Do not disconnect
equipment unless area is known to be nonhazardous.
Do not reset circuit breaker unless power
has been removed from the equipment or
the area is known to be non-hazardous.
Notice
• This manual should be passed on to the end user.
• The contents of this manual are subject to
change without prior notice.
• The contents of this manual shall not be
reproduced or copied, in part or in whole,
without permission.
• This manual explains the functions contained
in this product, but does not warrant that they
are suitable the particular purpose of the user.
• Every effort has been made to ensure
accuracy in the preparation of this manual.
However, when you realize mistaken
expressions or omissions, please contact the
nearest Yokogawa Electric representative or
sales office.
• This manual does not cover the special
specifications. This manual may be left
unchanged on any change of specification,
construction or parts when the change does
not affect the functions or performance of
the product.
• If the product is not used in a manner
specified in this manual, the safety of this
product may be impaired.
Yokogawa is not responsible for damage to the
instrument, poor performance of the instrument
or losses resulting from such, if the problems
are caused by:
• Improper operation by the user.
• Use of the instrument in improper
applications
• Use of the instrument in an improper
environment or improper utility program
• Repair or modification of the related
instrument by an engineer not authorized by
Yokogawa.
Safety and Modification Precautions
• Follow the safety precautions in this manual
when using the product to ensure protection
and safety of the human body, the product
and the system containing the product.
How to dispose the batteries:
This is an explanation about the new EU
Battery Directive (DIRECTIVE 2006/66/
EC). This directive is only valid in the EU.
Batteries are included in this product. Batteries
incorporated into this product cannot be
removed by yourself. Dispose them together
with this product. When you dispose this
product in the EU, contact your local Yokogawa
Europe B.V.office. Do not dispose them as
domestic household waste.
Battery type: silver oxide battery
Notice:
The symbol (see above) means they shall be
sorted out and collected as ordained in ANNEX
II in DIRECTIVE 2006/66/EC.
Product Disposal
The instrument should be disposed of in
accordance with local and national legislation/
regulations.
ii
The following safety symbols are used
on the product as well as in this manual.
DANGER
This symbol indicates that an operator
must follow the instructions laid out in
this manual in order to avoid the risks, for
the human body, of injury, electric shock,
or fatalities. The manual describes what
special care the operator must take to
avoid such risks.
WARNING
This symbol indicates that the operator
must refer to the instructions in this
manual in order to prevent the instrument
(hardware) or software from being
damaged, or a system failure from
occurring.
CAUTION
This symbol gives information essential
for understanding the operations and
functions.
Note!
This symbol indicates information that
complements the present topic.
This symbol indicates Protective Ground
Terminal
This symbol indicates Function Ground
Terminal (Do not use this terminal as the
protective ground terminal.)
This symbol indicates Alternating current.
This symbol indicates Direct current.
Warranty and service
Yokogawa products and parts are
guaranteed free from defects in
workmanship and material under normal
use and service for a period of (typically)
12 months from the date of shipment
from the manufacturer. Individual sales
organisations can deviate from the typical
warranty period, and the conditions of
sale relating to the original purchase order
should be consulted. Damage caused by
wear and tear, inadequate maintenance,
corrosion, or by the effects of chemical
processes are excluded from this warranty
coverage.
In the event of warranty claim, the defective
goods should be sent (freight paid) to the
service department of the relevant sales
organisation for repair or replacement
(at Yokogawa discretion). The following
information must be included in the letter
accompanying the returned goods:
• Part number, model code and serial
number
• Original purchase order and date
• Length of time in service and a
description of the process
• Description of the fault, and the
circumstances of failure
• Process/environmental conditions that
may be related to the failure of the
device.
• A statement whether warranty or nonwarranty service is requested
• Complete shipping and billing
instructions for return of material, plus
the name and phone number of a
contact person who can be reached for
further information.
Returned goods that have been in contact
with process fluids must be decontaminated/
disinfected before shipment. Goods should
carry a certificate to this effect, for the health
and safety of our employees. Material safety
data sheets should also be included for all
components of the processes to which the
equipment has been exposed.
Authorized Representative in EEA
The Authorized Representative for this
product in EEA is Yokogawa Europe B.V.
(Euroweg 2, 3825 HD Amersfoort, The
Netherlands).
1
The Yokogawa SC450G is a converter designed for
industrial process monitoring, measurement and
control applications. This instruction manual contains
the information needed to install, set up, operate and
maintain the unit correctly. This manual also includes
a basic troubleshooting guide to answer typical user
questions.
Yokogawa can not be responsible for the
performance of the SC450G converter if these
instructions are not followed.
1-1. Instrument check
Upon delivery, unpack the instrument carefully
and inspect it to ensure that it was not
damaged during shipment. If damage is found,
retain the original packing materials (including
the outer box) and then immediately notify the
carrier and the relevant Yokogawa sales office.
Make sure the model number on the nameplate
affixed to the top of the instrument agrees with your
order. Example of the nameplate is shown below.
1-2. Application
The SC450G converter is intended to be used for
continuous on-line measurement of Conductivity,
Resistivity and/or Concentration in industrial
installations. The unit combines simple operation
and microprocessor-based performance with
advanced self-diagnostics and enhanced
communications capability to meet the most
advanced requirements. The measurement can
be used as part of an automated process control
system. It can also be used to indicate operating
limits of a process, to monitor product quality, or
to function as a controller for a dosing/dilution
system.
Sensors should normally be mounted close to
the converter in order to ensure easy calibration
and peak performance. If the unit must be
mounted remotely from the sensors, WF10
extension cable can be used, up to a maximum
of 50 metres (150 feet), with a BA10 junction
box, and up 10 metres standard sensor cable.
The SC450G is delivered with a general
purpose default setting for programmable items
(see Chapter 5). While this initial configuration
allows easy start-up, the configuration should
be adjusted to suit each particular application.
An example of an adjustable item is the type
of temperature sensor used. The SC450G can
be adjusted for a number of different types of
temperature sensors.
Details provided in this instruction manual
are sufficient to operate the SC450G with all
Yokogawa sensor systems and a wide range of
third-party commercially available probes. For
best results, read this manual in conjunction
with the corresponding sensor instruction
manual.
Figure 1-1. Nameplate
Note!
• The nameplate will also contain the serial
number and any relevant certification marks.
Be sure to apply correct power to the unit, as
detailed on the nameplate.
• Refer to “Control drawing for FM approval” of
Appendix 4 in case of Option code is “-U (FM
version)” of Model and Suffix Codes.
• When contact output current is more than 4
Amps. ambient temperature should be less
than 40ºC.
Yokogawa designed the SC450G converter to
withstand industrial environments. It meets all
the CE regulatory standards. The unit meets or
exceeds stringent requirements (see section
2) without compromise, to assure the user of
continued accurate performance in even the
most demanding industrial installations.
IM 12D08N05-01E
1 INTRODUCTION AND GENERAL DESCRIPTION
1. INTRODUCTION AND GENERAL DESCRIPTION
2
2. GENERAL SPECIFICATIONS OF EXAxt SC450G
A) Inputs specifications : Two or four electrodes measurement with square wave excitation, using
max 60 m (200 ft) cable (WU40/WF10) and cell constants from 0.005 to
50.0 cm-1
B) Input ranges
Conductivity
: 0.000 µS/cm - 2000 mS/cm
Minimum
: 1 µS (underrange 0.00 µS x c.c.)
Maximum
: 200 mS (overrange 2000 mS x c.c.)
Resistivity
Minimum
Maximum
: 0.0 Ω•cm - 1000 MΩ•cm
: 5 Ω / c.c. (underrange 0.0 Ω/c.c.)
: 1 MΩ / c.c. (overrange 1000 MΩ/c.c.)
Temperature
Pt1000
: -20 to 250 ºC (0 - 500ºF)
Pt100
: -20 to 200 ºC (0 - 400ºF)
Ni100
: -20 to 200 ºC (0 - 400ºF)
NTC 8k55
: -10 to 120 ºC (10 - 250ºF)
Pb36 (JIS NTC 6k) : -20 to 120 ºC (0 - 250ºF)
C) Accuracy
Conductivity/resistivity : ≤ 0.5 % of reading
Temperature
: ≤ 0.3 ºC (≤ 0.4 ºC for Pt100)
Step response
: ≤ 4 sec for 90% (for a 2 decade step)
Note on performance specifications
The following tolerance is added to above performance.
mA output tolerance: ±0.02 mA of “4-20 mA“
D) Transmission signals
General
: Two isolated outputs of 4-20 mA. DC with common negative
Maximum load 600Ω. Bi-directional HART® digital communication,
superimposed on mA1 (4-20 mA) signal
Output function
: Linear or non-linear 21-step table for Conductivity/Resistivity,
concentration or temperature
Control function
: PID control
Burnout function : Burn up (21.0 mA) or burn down (3.6 mA) to signal failure acc. NAMUR
NE43
: Adjustable damping
: Expire time
Hold
: The mA-outputs are frozen to the last/fixed value during calibration/
commissioning
E) Contact outputs
General
: Four SPDT relay contacts with display indicators
Switch capacity
: Maximum values 100 VA, 250 V AC, 5 Amps. (*)
Maximum values 50 Watts, 250 V DC, 5 Amps. (*)
Status
: High/Low process alarms, selected from conductivity, resistivity,
concentration or temperature. Configurable delay time and hysteresis
: PID duty cycle or pulsed frequency control
: FAIL alarm
Control function
: On / Off
: Adjustable damping
: Expire time
IM 12D08N05-01E
3
IM 12D08N05-01E
2 GENERAL SPECIFICATIONS
Hold
: Contact can be used to signal the HOLD situation.
Fail safe
: Contact S4 is programmed as a fail-safe contact
(*)Note: When contact output current is more than 4 Amps, ambient temperature should be less
than 40 ºC.
F) Contact input
: Remote range switching to 10 times the programmed range.
Contact open
: If impedance > 100 kΩ: 1 x Range
(When programmed range for mA output is conductivity)
Contact closed
: If impedance < 10 Ω: 10 x Range
G) Temperature compensation
Function
: Automatic or manual, for temperature ranges mentioned under B
(inputs).
Reference temp. : programmable from 0 to 100 ºC or 30 - 210 ºF (default 25 ºC).
H) Compensation algorithm
: According IEC 60746-3 NaCl tables (default).
Two independent user programmable temperature coefficients, from 0%
to 3.5% per ºC (ºF) by adjustment or calibration.
Matrix compensation: With conductivity function of concentration and temperature. Choice out
of 13 preprogrammed matrixes and 2 100-points user-programmable
matrixes.
I) Calibration
: Semi-automatic calibration using pre-configured OIML (KCl) buffer
tables, with automatic stability check. Manual adjustment to grab sample
J) Logbook
: Software record of important events and diagnostic data readily available
in the display or through HART®.
K) Display
: Graphical Quarter VGA (320 x 240 pixels) LCD with LED backlight and
touchscreen. Plain language messages in English, German, French,
Spanish, Italian, Swedish, Portuguese and Japanese.
L) Form
Dimension
: 144 (W) x 144 (H) x Approx.144 (D) mm
Converter weight : Approx. 1.5 kg
M) Housing
: Cast aluminum housing with chemically resistant coating;
Polycarbonate cover with Polycarbonate flexible window
:
Protection IP66 / NEMA 4X / CSA Type 3S
Colour
: Silver grey
SC450G-A(D)-A
: IP66 cable glands are supplied with the unit
SC450G-A(D)-U : NEMA 4X close up plugs are mounted in the unused cable entry holes
and can be replaced by conduit fittings as required
Pipe, Panel or Wall mounting using optional hardware
Optional conduit adapter
G1/2, 1/2NPT or M20 female
N) Power supply
SC450G-A
: Ratings; 100-240 V AC Acceptable range; 90 to 264 V AC
Ratings; 50/60 Hz
Acceptable range; 50 Hz ±5%, 60 Hz ±5%
Power Consumption; 15 VA
SC450G-D
: Ratings; 12-24 V DC
Acceptable range; 10.8 to 26.4 V DC
Power Consumption; 10 W
O) Safety and EMC conforming standards
Safety
: EN 61010-1
EN 61010-2-030
EN 61010-2-201
CSA C22.2 No.61010-1
UL 61010-1
4
EMC
: EN 61326-1 Class A, Table 2 (*1)
EN 61326-2-3
EN 61000-3-2 Class A
EN 61000-3-3
RCM: EN 61326-1 Class A
Korea Electromagnetic Conformity Standard
FM nonincendive approval (suffix code Type: -U):
FM3611 Class I, Div.2, Group ABCD,T6 for Ta -20 to 55ºC
*1: Influence of immunity environment (Criteria A): Output shift is specified within ±35% of F.S.
Installation altitude: 2000 m or less
Category based on IEC 61010: II (Note)
Pollution degree based on IEC 61010: 2 (Note)
Note: Installation category, called over-voltage category, specifies impulse withstand voltage.
Category II is for electrical equipment.
Pollution degree indicates the degree of existence of solid, liquid, gas or other
inclusions which may reduce dielectric strength. Degree 2 is the normal indoor
environment.
CAUTION
This instrument is a Class A product, and it is designed for use in the industrial environment.
Please use this instrument in the industrial environment only.
P) Environment and operational conditions
Ambient temperature : -20 to +55 ºC (-5 - 130 ºF)
Storage temperature : -30 to +70 ºC (-20 - 160 ºF)
Humidity
: 10 to 90% RH at 40 ºC (100 ºF) (non-condensing)
Data protection
: EEPROM for configuration data and logbook. Lithium cell for clock
Watchdog timer
: Checks microprocessor
Power down
: Reset to measurement
Automatic safeguard : Auto return to measuring mode when touchscreen is untouched for 10 min.
Model and Suffix Codes
Model
[Style: S2]
Option code
Description
SC450G
-----------------
-----------------
Conductivity/Resistivity Converter
Power
-A
-D
---------------------------------
AC version (100…240 VAC)
DC version (12…24 VDC)
---------------------------------
General purpose version
FM version (*3)
Type
-A
-U
Mounting Hardware
Hood
Conduit adapter
Tag Plate
/UM
/U
/PM
/H5
/AFTG
/ANSI
/AM20
/SCT
Universal mounting kit (panel, pipe, wall)
Pipe and wall mounting hardware (*2)
Panel mounting hardware (*2)
Awning hood (stainless steel) (*2)
G1/2 (*2)
1/2NPT (*2)
M20 (*2)
Stainless steel tag plate (*1)
*1: If the tag number is predefined with the purchase, Yokogawa will inscript the tag plate with the specified tag number, and
program the tag number in the converter.
*2: Option codes /U, /PM, /H5, /AFTG, /ANSI and /AM20 are not specified for FM version (-U).
*3: Nonincendive for FM only
IM 12D08N05-01E
5
3. INSTALLATION AND WIRING
3-1-2. Mounting methods
Refer to figures 3-2 and 3-3. Note that the
SC450G converter has universal mounting
capabilities:
• Panel mounting using optional brackets
• Surface mounting on a plate (using bolts
from the back)
• Wall mounting on a bracket (for example,
on a solid wall)
• Pipe mounting using a bracket on a
horizontal or vertical pipe
Size nominal 50A
Unit: mm (inch)
Hood (Option code: /H5, some cutout on the left side cover)
184(7.2")
220(8.66")
144(5.67")
4x M6 screw
depth 7
72(2.8")
24.5(1")
144(5.67")
20
(0.79")
M20 cable gland
(When shipped, not installed)
20
(0.79")
121.5(4.78")
A
D
B
C
E
F
Adapter for Conduit Work (optional)
(option code : / AFTG, / ANSI, / AM20)
A : For output signal
Adapter
B : For contact input
C : For sensor cable
D : For contact output (S1 and S2)
E : For contact output (S3 and S4)
F : For power supply
Grounding terminal (M4 screw)
Approx.
49
(1.93") 55(2.2")
G1/2 screw (/AFTG), 1/2 NPT screw (/ANSI)
M20 screw (/AM20)
Figure 3-1. Housing dimensions and layout of glands
IM 12D08N05-01E
3 INSTALLATION AND WIRING
3-1. Installation and dimensions
3-1-1. Installation site
The SC450G converter is weatherproof and
can be installed inside or outside. It should,
however, be installed as close as possible to
the sensor to avoid long cable runs between
sensor and converter. In any case, the cable
length should not exceed 60 metres (197 feet).
Select an installation site where:
• Mechanical vibrations and shocks are
negligible
• No relay/power switches are in the direct
environment
• Access is possible to the cable glands
(see figure 3-1)
• The converter is not mounted in direct
sunlight or severe weather conditions
• Maintenance procedures are possible
(avoiding corrosive environments)
The ambient temperature and humidity of the
installation environment must be within the
limits of the instrument specifications. (See
chapter 2).
6
Unit: mm (inch)
+1
138 0
(5.43")
min.195(7.75")
min.185 (7.25")
+1
138 0
(5.43")
138(5.43")
M6
M5
138
(5.43")
M6
Figure 3-2. Option /PM: panel mounting diagram
(Note) When option code "/UM" is specified, universal pipe/wall/panel mounting kit are supplied--same as option code "/U" and "/PM" both specified.
wall mounting
13
(0.51")
80
(3.15")
pipe mounting
(vertical)
pipe mounting
(horizontal)
4x Ф6.5
(0.26")
80
(3.15")
200
(7.87")
4x Ф10
(0.4")
35
(1.38")
15
(0.6")
70
(2.75")
154.5
(6.08")
2" ND. pipe
Figure 3-3. Option /U: wall and pipe mounting diagram
IM 12D08N05-01E
7
WARNING This connector for software must be used only by
connector for (future) software
Yokogawa’s service personnel.
input
terminal
block
potentiomerter
LCD
bracket
protective
shield
bracket
6 X M20 glands
Note: SC450G-A(D)-U
The enclosure is provided with stoppers in stead of M20 cable glands for the unused holes.
These stoppers must be removed and replaced by FM approved conduit fittings in accordance with good installation practice. Also see Appendix 7, Control drawing for FM approval.
Figure 3-4. Internal view of SC450G wiring compartment
3-2. Wiring
3-2-1. Preparation
Refer to figure 3-4. The relay contact terminals
and power supply connections are under the
screening (shielding) plate. These should be
connected first. Connect the sensor, outputs
and HART® communication connections last.
To open the SC450G for wiring:
1. Loosen the four frontplate screws and swing
open the cover.
2. The upper terminal strip is now visible.
3. Remove the screen (shield) plate covering
the lower terminal strip.
4. Connect the power supply and contact
outputs. Use the three glands at the back for
these cables.
DANGER
• Cables that withstand temperatures of at
least 75 °C should be used for wiring.
• Wiring work should be performed to meet
IP66 or higher requirements. Tighten four
frontplate screws to 1.5 N·m torque.
• Use cables that comply with UL2556VW-1 or equivalent.
WARNING
Always place the screen plate over the power
supply and contact terminals for safety reasons
and to avoid interference.
5. Put back (replace) the screen (shield) plate
over the lower terminals.
6. Connect the analog output(s), the sensor
inputs, and, if necessary, the HART® wiring
and input contact.
7. Use the front three glands for analog output,
sensor inputs, contact input and HART®
wiring (see figure 3-5).
8. Swing back the cover and secure it with the
four screws.
9. Switch on the power. Commission the
instrument as required or use the default
settings.
CAUTION
Do not turn on power with the touchcsreen
pressed, otherwise inaccurate screen
positioning will occur. If it occurs, leave the
touchscreen unpressed, turn off power then
on again. The screen positioning will be
accurate.
IM 12D08N05-01E
3 INSTALLATION AND WIRING
output
terminal
block
10
3-3-3. AC power
Connect terminal L to the phase line of the AC
power and terminal N to the zero line. See figure
3-8 for the power ground. This is separated from
input ground by a galvanic isolation.
Should be used the cable beyond 300 V AC.
WARNING
Fuse replacement should be performed only by
a qualified service personnel.
See Sec.7. MAINTENANCE, Fuse
Fuse ratings:
Power supply
Fuse type
12-24VDC, 10 W max 2 A/250 V, Slow
100-240 V AC, 15 V A max
0.5 A/250 V, Slow
3-3-4. DC power
Connect terminal 1 to the positive outlet and
terminal 2 to the negative outlet.
Should be used the cable beyond 30 V AC.
This is separated from input ground by a
galvanic isolation. The size of conductors
should be at least 1.25 mm2. The overall cable
diameter should be between 6 & 12 mm.
3-3-2. Access to terminal and cable entry
Terminals 1 and 2 are used for the power
supply. Guide the power cables through the
gland closest to the power supply terminals. The
terminals will accept wires of 2.5 mm2 (14 AWG).
Always use cable finishings if possible.
2
1
-
2
+
1
N
POWER
L
POWER
100-240 VAC/15 VA/ 50/60Hz
FUSE: 500 mA/250 VAC/T
12-24 V
/10 W
FUSE: 2A/250 VAC/T
AC
DC
32
31
NC
33
C
42
NO
NC
S1
63
SHLD
66
-
41
C
43
NO
S2
65
+
mA2
62
61
+
mA1
22
250V / 5A
52
AC / DC
100VA / 50W NC
CONTACTS
21
11
+
(+HART)
12
-
51
C
53
72
NO
NO
S3
71
C
73
NC
S4 (fail-safe)
13
14
15
16
INNER OUTER INNER OUTER
TEMP
CONTACT
SENSOR(S)
mA OUTPUTS
REFER TO INSTRUCTION MANUAL FOR CONNECTIONS
ELECTRODE
SC
Figure 3-7. Input and output connections
3-3-5. Grounding the housing
For the safety of the user and to protect the
instrument against interference, the housing
must always be connected to ground. This has
to be done by a large area conductor. This
cable can be fixed to the rear of the housing or
by using the internal ground connections using
a braided wire cable. See figure 3-8.
DANGER
The minimum cross sectional area of the
protective grounding wire should be 0.75 mm2.
IM 12D08N05-01E
3-3-6. Switching on the instrument
After all connections are made and checked,
the power can be switched on from the power
supply. Make sure the LCD display comes on.
After a brief interval, the display will change to
the measured value. If errors are displayed or
a valid measured value is not shown, consult
the troubleshooting section (Chapter 8) before
calling Yokogawa.
11
(M4 screw)
2
1
L
POWER
100-240 VAC/15 VA/ 50/60Hz
FUSE: 500 mA/250 VAC/T
AC
(M4 screw)
Figure 3-8-a. External grounding
3-4. Wiring the contact signals
3-4-1. General precautions
The contact output signals consist of voltagefree relay contacts for switching electrical
appliances (SPDT). They can also be used as
digital outputs to signal processing equipment
(such as a controller or PLC). It is possible to
use multi-core cables for the contact in and
output signals and shielded multi-core cable for
the analog signals.
3-4-2. Contact outputs.
The SC450G unit’s four contacts (switches)
that can be wired and configured to suit user
requirements. Contact S4 is programmed as
a fail-safe contact. Please refer to section
5-7, Contact output setup for functionality
description.
Alarm (limits monitoring)
Contacts configured as “ALARM” can be
energized when limits are crossed.
Fail
Contacts configured as “FAIL” will be
energized when a fail situation occurs. Some
fail situations are automatically signaled by
the internal diagnostics (electronics) of the
converter. Others can be configured by the
user (see section 5-12 Error Configuration). By
pressing the “INFO” button on the main screen
the user is given an explanation as well as a
remedy for the current fail situation.
Always connect the fail contact to an alarm
device such as a warning light, alarm bell or
displayed on an annunciator.
Figure 3-8-b. Internal grounding
“ALARM” Contact “FAIL” Contact
Power Off
NC
NC
Power On
NC
NC
Alarm
NO
NC
Fail
NC
NO
Fail and Alarm
NC*
NO
HOLD
NC
NC
* When a fail situation occurs which is related
to the parameter associated with the contact
(Conductivity, Concentration or temperature) the
contact will go to NC. When the fail situation is not
related to the parameter associated with the contact
the contact will remain in the state it is currently in.
Be beyond 300 V AC and please use a cable
of the electric current capacity according to the
load connected to contact output.
3-4-3. Contact input
It is necessary to use screening/shielding on
the input signal cables. Terminal 63 is used to
connect the shielding.
3-5. Wiring the mA-output signals
3-5-1. General precautions
The analog output signals of the SC450G
transmit low power standard industry signals to
peripherals like control systems or strip-chart
recorders (Figure 3-6).
3-5-2. Analog output signals
The output signals consist of active current
signals of 4-20 mA. The maximum load can be
600 ohms on each.
Should be used the cable beyond 30 V AC.
It is necessary to use screening/shielding on
the output signal cables. Terminal 63 is used to
connect the shielding.
IM 12D08N05-01E
3 INSTALLATION AND WIRING
N
20
5. MENU STRUCTURE COMMISSIONING
Measurement setup
Main parameter
Choose the required parameter, either conductivity or resistivity. If the main parameter is
changed the instrument will reset main display
settings, units and recalculate several values.
The menu structure will change accordingly.
5-1. Configure sensor
Sensor type
Choose the sensor type used. Normally
conductivity and/or resistivity measurements
are done with 2-electrode type sensors. At
high conductivity ranges, polarization of the
electrodes may cause an error in conductivity
measurement. For this reason 4-electrode type
sensors may be necessary.
Measuring unit /cm /m
Either /cm or /m can be chosen here. The process values will be expressed in S/cm or S/m
respectively, (Ω.cm or Ω.m in resistivity mode).
Note: When measuring unit is changed,
parameter of “mA output setup” and
“contact output setup” will be cleared.
Set these parameter again refering to
subsection 5-6 to 5-7.
If you replace the sensor, delete the predictive
maintenance log first on the logbook configuration screen in section 5-12. If you do not,
proper prediction will not be possible using the
sensor diagnostics.
You can check whether the log has been
properly cleared by changing the sensor data
to predictive maintenance under Read logbook
on the Detail screen of figure 4.3 and then
pressing Enter.
Cell constant (factory)
Cell constant given by factory calibration.
Usually given on a label on the sensor or the
calibration certificate. Only change this value
in case a new sensor is used. By changing this
value the actual cell constant is also changed.
Measure
Process values to be measured can be selected to suit the user’s preference.: Conductivity only, Concentration only or one of both
Conductivity and Concentration.
Note: this choice is not available in
Resistivity mode.
IM 12D08N05-01E
5-2. Temperature setting
Temperature Element
Selection of the temperature sensor used for
compensation. The default selection is the
Pt1000 Ohm sensor, which gives excellent
precision with the two wire connections used.
The other options give the flexibility to use a
very wide range of other conductivity/resistivity
sensors.
Temperature Unit
Celcius or Fahrenheit temperature scales can
be selected to suit the user’s preference.
When the unit is changed all temperature
related parameters and settings will be recalculated.
5-3. Temperature compensation
Compensation
Two types of methods can be used here. Automatic for use of temperature element. Select
one of the Temperature elements used. The
other is a manual set temperature. The manual
temperature that represents the process temperature must be set here.
Reference Temperature
Choose a temperature to which the measured conductivity (or resistivity) value must be
compensated. Normally 25°C (77ºF) is used,
therefore this temperature is chosen as the
default value.
Method
TC In addition to the temperature coefficient
calibration routine it is possible to adjust the
compensation factor directly. If the compensation factor of the sample liquid is known from
laboratory experiments or has been previously
determined, it can be introduced here.
Adjust the value between 0.00 to 3.50 % per
°C. In combination with reference temperature
a linear compensation function is obtained,
suitable for all kinds of chemical solutions.
NaCl Temperature compensation according
NaCl curve. See appendix 1 for values.
Matrix The SC450G is equipped with a matrix
type algorithm for accurate temperature
compensation in various applications. Select
the range as close as possible to the actual
temperature/concentration range. The SC450G
will compensate by interpolation. If user defined
41
9. QUALITY INSPECTION
Quality
Inspection
Standards
1.
SC450G
Conductivity or Resistivity Converter
Scope
This inspection standard applies to the SC450G Conductivity or Resistivity Converter.
2.
Inspection Items
2.1
*2.2
*2.3
2.4
2.5
Insulation resistance test
Dielectric strength test
Sensor signal input test
Temperature indication check
Current output test
Note: Items marked with an asterisk (*) may only be confirmed by a test certificate.
3.
Inspection Methods, Standards and Conditions
3.1
Connect the testing circuit as shown in Figure 1. Allow the instrument to warm up for
at least 5 minutes before conducting the tests. For the connections for the insulation
resistance and dielectric strength tests, follow the instructions in Sections 3.1 and 3.2.
Insulation Resistance Test
Apply 500 V DC between the terminals shown below. The insulation resistance must be
100 MΩ or greater.
(1) Between the power supply terminals shorted together (1 and 2) and the protective
earth terminal ( )
(2) Between the contact output terminals shorted together (31, 32, 33, 41, 42, 43, 51, 52,
53, 71, 72 and 73) and the protective earth terminal ( )
(3) Between the current output terminals shorted (61, 62, 65 and 66) and the protective
earth terminal ( )
3.2
Dielectric Strength Test
3.3
Sensor Signal Input Test
Connect the testing circuit as shown in Figure 1 and set the equipment as follows:
Decade resistance box 1 (temperature simulation input):
1097.3 [Ω]
Decade resistance box 2 (conductivity simulation input):
150 [Ω]
The power supply voltage should be set in accordance with the specifications of the
converter.
QIS 12D08N05-01E
1st Edition: Jul. 2007
2nd Edition: Sep. 2013
IM 12D08N05-01E
9 QUALITY INSPECTION
(1) Apply 1400 V AC, an AC voltage of substantially sinusoidal waveform with a frequency
of 50 Hz or 60 Hz, between the power supply terminals shorted together (1 and 2) and
the protective earth terminal ( ), for at least 2 seconds. The insulation must withstand
this voltage. (The sensed current should be 10 mA.)
(2) Apply 1400 V AC, an AC voltage of substantially sinusoidal waveform with a frequency
of 50 Hz or 60 Hz, between the contact output terminals shorted together (31, 32, 33,
41, 42, 43, 51, 52, 53, 71, 72 and 73) and the protective earth terminal ( ), for at
least 2 seconds. The insulation must withstand this voltage. (The sensed current
should be 10 mA.)
(3) Apply 500 V AC, an AC voltage of substantially sinusoidal waveform with a frequency
of 50 Hz or 60 Hz, between the current output terminals shorted (61, 62, 65 and 66)
and the protective earth terminal ( ), for at least 2 seconds. The insulation must
withstand this voltage. (The sensed current should be 10 mA.)
42
2/3
This test is done on the “HIF” display of “Factory Mode”.
a. Touch the [Setup] icon.
b. Touch the [Commissioning].
c. Touch the [Advanced setup].
d. Touch the [Factory adjustment].
e. Enter a password.
f. Select the [Factory Mode] in “Key.”
g. Select the [HIF] in “Execute.”
When the resistance of the decade resistance box 2 to the corresponding value
“RANGE1” to “RANGE5” in Table 1 is set, check the data display and the value must be
within the range shown in Table 1.
Table 1
RANGE
RANGE1
RANGE2
RANGE3
RANGE4
RANGE5
Resistance (Ω) of
Decade Resistance Box 2
20
200
2k
20k
200k
Data Display (Ω)
20
200
2k
20k
200k
±0.1
±1
±0.01k
±0.1k
±1k
After the above test is completed, touch the [Exit] to return to the “HIF” display.
3.4
Temperature Indication Check
Following Section 3.3, select the [PT1000] in “Temperature” of the “HIF” display.
In this state, change the resistance of the decade resistance box 1 and check the data
display. The value on the data display must be within the range shown in Table 2.
Table 2
Temperature (°C)
–10
25
240
Resistance () of
Decade Resistance Box 1
960.9
1097.3
1904.6
Data Display (°C)
–10 ±0.3
25 ±0.3
240 ±0.3
After the above test is completed, touch the [Exit] to return to the “HIF” display.
3.5
Current Output Test
Following Section 3.4, select the [Check] in “mA outputs” of the “HIF” display. “Set value
4.000 mA” appears at the bottom of the display. Select “Next value” in the “Command”
and touch “Enter,” the value on the data display increases in steps of 4 mA. Check the
current outputs 1 and 2 corresponding to the data display, the current output must be
within the range shown in Table 3.
Table 3
Data Display
4
12
20
Current Output (mA DC)
4 ±0.02
12 ±0.02
20 ±0.02
QIS 12D08N05-01E
IM 12D08N05-01E
43
3/3
After all tests are completed,
a. Touch the [Exit] twice to return to the “Service” display.
b. Select “Normal” in “Key”.
c. Touch the [Home] icon to return to the initial display.
Figure 1 Testing Circuit and Test Equipment
9 QUALITY INSPECTION
QIS 12D08N05-01E
IM 12D08N05-01E
44
成 績 表
TEST CERTIFICATE
製品名称
4線式導電率変換器
PRODUCT NAME 4 WIRE CONDUCTIVITY CONVERTER
形名
MODEL
SC450G
手配No.
ORDER NO.
タグNo.
TAG NO.
計器番号
SERIAL NO.
検 査 項 目 INSPECTION ITEM
外 観
APPEARANCE
絶縁抵抗
INSULATION
RESISTANCE
耐電圧
DIELECTRIC
STRENGTH
センサ入力表示
SENSOR
INPUT
INDICATION
温度表示
TEMPERATURE
INDICATION
出力電流
CURRENT
OUTPUT
結 果 RESULT
電源端子(1,2)一括と保護接地端子間
接点出力端子(31,32,33,41,42,43,51,52,53,71,72,73)一括と保護接地端子間
電流出力端子(61,62,65,66)一括と保護接地端子間
100MΩ以上 / 500V DC
BETWEEN POWER SUPPLY TERMINALS (1,2) AND PROTECTIVE EARTH TERMINAL
BETWEEN CONTACTS TERMINALS (31,32,33,41,42,43,51,52,53,71,72,73) AND PROTECTIVE EARTH TERMINAL
BETWEEN CURRENT OUTPUT TERMINALS (61,62,65,66) AND PROTECTIVE EARTH TERMINAL
100MΩ OR MORE / 500V DC
電源端子(1,2)一括と保護接地端子間
接点出力端子(31,32,33,41,42,43,51,52,53,71,72,73)一括と保護接地端子間
1400V AC /2秒間
BETWEEN POWER SUPPLY TERMINALS (1,2) AND PROTECTIVE EARTH TERMINAL
BETWEEN CONTACTS TERMINALS (31,32,33,41,42,43,51,52,53,71,72,73) AND PROTECTIVE EARTH TERMINAL
1400V AC /2sec
電流出力端子(61,62,65,66)と保護接地端子間
500V AC /2秒間
BETWEEN CURRENT OUTPUT TERMINALS (61,62,65,66) AND PROTECTIVE EARTH TERMINAL
500V AC /2sec
抵抗値
レンジ
RANGE
RESISTANCE
RANGE1
RANGE2
RANGE3
RENGE4
RANGE5
表示 INDICATION (Ω)
実測値
(Ω)
基準値
許容差
REFERENCE
ACCURACY
20
200
2K
20K
200K
20
200
2K
20K
200K
±0.1
±1
±0.01K
±0.1K
±1K
誤差
ACTUAL
ERROR
PT1000
抵抗値
RESISTANCE
基準値
表 示 INDICATION (℃)
許容差
実測値
(Ω)
REFERENCE
ACCURACY
960.9
1097.3
1904.6
-10
25
240
±0.3
±0.3
±0.3
許容差 : ±0.02mA DC
ACCURACY : ±0.02mA DC
出力1 OUTPUT1 (mA DC)
表示
実測値
誤差
基準値
INDICATION
REFERENCE
4
12
20
ACTUAL
ERROR
誤差
ACTUAL
ERROR
出力2 OUTPUT2 (mA DC)
基準値
実測値
REFERENCE
ACTUAL
誤差
ERROR
4
12
20
4
12
20
NOTES
日付
DATE
検査者
INSPECTOR
室内温度・湿度
AMBIENT TEMP. & HUM.
承認者
APPROVED BY
℃ &
%
QIC-12D08N05-01
Ed2: Sep. 2013
IM 12D08N05-01E
Customer
Maintenance
Parts List
Model SC450G [Style: S2]
Conductivity / Resistivity
Converter
7, 10
3
9
2
11
5c
5a, 5b
8
6a, 6b
12
1a, 1b
Item
*1a
*1b
*2
*3
Part No.
K9676GA
K9676HA
K9677EB
K9676MA
*5a
*5b
*5c
*6a
*6b
*7
8
9
10
*11
12
K9676MX
K9676MW
K9676MY
A1108EF
A1111EF
K9676BE
K9676BU
K9676DL
K9676BT
K9676CM
—
K9171SU
K9316AF
K9676BC
Qty
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
Description
Power board assembly AC version
Power board assembly DC version
Main board assembly SC version
LCD module
Cable assembly (3 core)
Cable assembly (10 core)
Cable assembly (shield)
Fuse AC version (1 pcs.)
Fuse DC version (1 pcs.)
Cover assembly without, screws and hingepins
Cable glands assembly (6 pcs. M20)
Stainless tagplate blank
Screw assembly to fix cover (M4 screws, washer, O-ring, hingepins)
Housing assembly polyurethan baked finish
Adapter assembly for conduit work
For G1/2 screw when /AFTG specified
For 1/2NPT screw when /ANSI specified
For M20 screw when /AM20 specified
*) Do not exchange these parts. Call service personnel.
©Copyright 2007, 4th Edition: Oct, 2011 (YK)
Subject to change without notice.
CMPL 12D08N05-02E
2
Pipe/Wall Mounting Hardware (Option code: /U)
Panel Mounting Hardware (Option code: /PM)
7
2
8
3
3
9
4
5
6
1
UNIVERSAL MOUNT SET (Option code : /UM) includes
both “/U” and “/PM”.
3
10
Awning Hood (Option code: /H5)
CMPL 12D08N05-02E
Item
1
2
3
4
5
Part No.
K9171SS
Y9608KU
D0117XL-A
K9171SY
K9171SX
Qty
1
4
1
1
1
Description
Mounting Set (/U)
Screw
U-Bolt Assembly
Plate
Bracket
6
7
8
9
K9171ST
Y9520LU
K9171SW
Y9608KU
1
2
2
4
Mounting Set (/PM)
Screw
Bracket
Screw
10
K9676BA
1
Awning hood assembly (/H5)
4th Edition : Oct. 2011 (YK)