Thermo FIsher 43iQ Instruction Manual
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
43iQ
Instruction Manual
Sulfur Dioxide Analyzer
117568-00 • 15Jan2018
Contents
Thermo Scientific
Chapter 1
Introduction ........................................................................................................ 1-1
iQ Series Instrument Platform............................................................. 1-1
43iQ Principle of Operation ............................................................... 1-3
Specifications ...................................................................................... 1-5
Dimensions ......................................................................................... 1-7
Chapter 2
Installation and Setup ...................................................................................... 2-1
Unpacking and Inspection .................................................................. 2-1
Cover Removing and Replacing .......................................................... 2-2
Mounting Options .............................................................................. 2-3
Bench Mount ................................................................................... 2-3
Rack Mount ..................................................................................... 2-4
Setup Procedure .................................................................................. 2-6
Startup ................................................................................................ 2-8
Chapter 3
Operation ............................................................................................................ 3-1
Instrument Display ............................................................................. 3-1
Main Menus and Keypads ................................................................ 3-5
Numeric Keypad ........................................................................... 3-6
Alphanumeric Keypad ................................................................... 3-7
Calibration .......................................................................................... 3-9
Calibrate Background..................................................................... 3-11
Calibrate Span Coefficient ............................................................. 3-12
Zero/Span Schedule ....................................................................... 3-14
Advanced Calibration ..................................................................... 3-16
Manual Calibration ..................................................................... 3-17
Calibration History ..................................................................... 3-22
Data .................................................................................................. 3-23
View Data Log (Last Hour)............................................................ 3-24
View Data Log (Last 24 Hours) ..................................................... 3-25
View Data Log (User Defined Time) ............................................. 3-26
Advanced Data Setup ..................................................................... 3-28
Data Logging Setup .................................................................... 3-29
Select Data Logging Variables ..................................................... 3-30
Streaming Data Setup ................................................................. 3-32
Select Streaming Variables ........................................................... 3-33
Settings ............................................................................................. 3-35
Health Check ................................................................................. 3-36
43iQ Instruction Manual
iii
Contents
Status and Alarms........................................................................ 3-37
Valve and Pump Resets ............................................................... 3-48
Predictive Diagnostics ................................................................. 3-54
Maintenance ............................................................................... 3-55
Preventive Maintenance .............................................................. 3-56
Change Part ................................................................................ 3-58
Maintenance History................................................................... 3-59
File Sharing and Support............................................................. 3-60
Measurement Settings .................................................................... 3-61
Averaging Time ........................................................................... 3-63
Range Mode Selection................................................................. 3-65
Range Setting .............................................................................. 3-67
Gas Mode ................................................................................... 3-68
Gas Units .................................................................................... 3-69
Advanced Measurement Settings ................................................. 3-70
Communications ........................................................................... 3-76
Wired TCP/DHCP .................................................................... 3-77
Serial RS-232/485 ....................................................................... 3-79
Analog I/O .................................................................................. 3-80
Digital I/O .................................................................................. 3-81
Email Server (SMTP) .................................................................. 3-82
Instrument Settings ........................................................................ 3-83
Display Setup .............................................................................. 3-84
Alarm Setpoints........................................................................... 3-85
Clock .......................................................................................... 3-87
Date / Time Parameters .............................................................. 3-88
Time Zone .................................................................................. 3-89
Time Server................................................................................. 3-91
Date Format................................................................................ 3-92
Configuration ................................................................................ 3-93
Security Access Levels ..................................................................... 3-94
Change Security to View Only Access ......................................... 3-96
Change Full Access Security Password ......................................... 3-97
USB Drive ..................................................................................... 3-99
Firmware Update Via USB Drive .............................................. 3-100
Download Data To USB Drive ................................................. 3-101
Change USB Password .............................................................. 3-104
User Contact Information ............................................................ 3-106
Update Bootloader ....................................................................... 3-107
Chapter 4
iv
43iQ Instruction Manual
Calibration .......................................................................................................... 4-1
Equipment Required ........................................................................... 4-1
Zero Air Generation ......................................................................... 4-2
Commercial Heatless Air Dryers ................................................... 4-2
Absorbing Column........................................................................ 4-2
SO2 Concentration Standard ............................................................ 4-2
Thermo Scientific
Contents
Calibration Gas Generation ............................................................. 4-2
Cylinder Gas Dilution................................................................... 4-3
External Flow Meter(s) and Controller(s) ...................................... 4-3
Commercial Precision Dilution Systems........................................ 4-4
Permeation Tube System............................................................... 4-4
Commercial Permeation Systems .................................................. 4-5
Calibration .......................................................................................... 4-6
Calibration in Dual/Auto Range Mode ............................................... 4-8
Zero and Span Check ........................................................................ 4-10
Manual Calibration ........................................................................... 4-11
Adjust Background......................................................................... 4-12
Adjust Span Coefficient.................................................................. 4-12
Reset Bkg to 0.000 and Span Coef to 1.000 ................................... 4-13
Zero/Span Schedule .......................................................................... 4-14
Next Time ..................................................................................... 4-14
Period ............................................................................................ 4-14
Zero/Span/Purge Duration Minutes .............................................. 4-14
Schedule Averaging Time ............................................................... 4-15
Background Calibration and Span Calibration ............................... 4-15
Zero/Span Ratio ............................................................................. 4-15
References ......................................................................................... 4-15
Thermo Scientific
Chapter 5
Maintenance ...................................................................................................... 5-1
Safety Precautions ............................................................................... 5-1
Fan Filter Inspection and Cleaning ..................................................... 5-1
Pump Rebuilding ................................................................................ 5-2
Leak Test ............................................................................................ 5-5
Lamp Voltage Check ........................................................................... 5-6
Lamp Voltage Adjustment................................................................... 5-6
Chapter 6
Troubleshooting ................................................................................................ 6-1
Safety Precautions ............................................................................... 6-1
Troubleshooting Guide ....................................................................... 6-1
Chapter 7
Servicing............................................................................................................. 7-1
Safety Precautions ............................................................................... 7-1
Firmware Updates ............................................................................... 7-3
Replacement Parts List ........................................................................ 7-3
Fuse Replacement ............................................................................... 7-5
Filter Replacement .............................................................................. 7-6
Fan Replacement................................................................................. 7-7
Measurement Side Removal and Replacing ......................................... 7-9
LCD Module Replacement ............................................................... 7-12
I/O Replacement ............................................................................... 7-14
43iQ Instruction Manual
v
Contents
Peripherals Support Board and System Controller Board
Replacement ..................................................................................... 7-16
DMC Pressure and Flow Board ........................................................ 7-17
Pump Replacement ........................................................................... 7-19
Capillary Cleaning and/or Replacement ............................................ 7-22
Capillary O-Ring Replacement ......................................................... 7-24
Power Supply Replacement ............................................................... 7-24
Step POL Board Replacement ........................................................... 7-26
DMC Optical Bench Removal .......................................................... 7-29
Optical Bench Assembly Removal .................................................. 7-29
Optical Bench Assembly Removal .................................................. 7-31
Photomultiplier Tube (PMT) Replacement ................................... 7-33
Flasher Pack Replacement .............................................................. 7-35
Flasher Pack Lamp Replacement .................................................... 7-36
Photo Lamp Detector Board Replacement ..................................... 7-37
Kicker Assembly Replacement ........................................................ 7-39
Optional Manifold Replacement ....................................................... 7-40
Optional DMC Permeation Oven Solenoid Valve Replacement ....... 7-42
Permeation Oven Replacement ......................................................... 7-44
Permeation Oven Board Replacement ............................................... 7-47
Chapter 8
vi
43iQ Instruction Manual
System Description ........................................................................................... 8-1
Optical Bench DMC........................................................................... 8-2
Heated Hydrocarbon Kicker ............................................................ 8-2
Optical Bench Hardware .................................................................. 8-2
Condensing Lens........................................................................... 8-2
Mirror Assembly ........................................................................... 8-2
Light Baffle ................................................................................... 8-2
Bandpass Filter .............................................................................. 8-2
Flash Lamp Trigger Assembly .......................................................... 8-3
Flash Lamp ................................................................................... 8-3
Flash Trigger Board....................................................................... 8-3
Flash Intensity Assembly ............................................................... 8-3
Photomultiplier Tube ...................................................................... 8-3
Optical Bench DMC Board ............................................................. 8-3
Bench Heater ................................................................................ 8-4
Permeation Oven (Optional) .............................................................. 8-4
Common Electronics .......................................................................... 8-4
Power Supply ................................................................................... 8-7
Front Panel ...................................................................................... 8-7
I/O and Communication Components ............................................ 8-7
System Controller Board .................................................................. 8-7
Backplane Board .............................................................................. 8-7
Peripherals Support System ................................................................. 8-8
Fan ................................................................................................... 8-8
STEP POL Board ............................................................................ 8-8
Thermo Scientific
Contents
Sample Pump ................................................................................... 8-8
Solenoid Valve Panel ........................................................................ 8-8
Flow/Pressure DMC ........................................................................... 8-8
Firmware ............................................................................................. 8-8
Thermo Scientific
Chapter 9
Optional Equipment .......................................................................................... 9-1
Connecting External Devices .............................................................. 9-1
Communication Board .................................................................... 9-2
RS-232/RS-485 Port ..................................................................... 9-2
RS-485 External Accessory Port .................................................... 9-3
Analog I/O Board ............................................................................ 9-4
Analog Voltage Inputs ................................................................... 9-4
Analog Voltage Outputs ................................................................ 9-5
Analog Output Calibration .............................................................. 9-6
Analog Output Zero Calibration ................................................... 9-7
Analog Output Full Scale Calibration ........................................... 9-9
Digital I/O Board........................................................................... 9-11
Digital Inputs .............................................................................. 9-11
Digital Relay Switches ................................................................. 9-13
Valve Driver Outputs .................................................................. 9-15
Internal Zero/Span and Sample Valves .............................................. 9-17
Internal Permeation Span Source ...................................................... 9-17
Permeation Tube Installation ......................................................... 9-19
Computation of Concentrations .................................................... 9-22
PTFE Particulate Filter ..................................................................... 9-23
Appendix A
Safety, Warranty, and WEEE .......................................................................... A-1
Safety .................................................................................................. A-1
Safety and Equipment Damage Alerts .............................................. A-1
Warranty ............................................................................................. A-2
WEEE Compliance ............................................................................. A-4
WEEE Symbol ................................................................................. A-4
Appendix B
Quick Reference .............................................................................................. B-1
List of Figures ..................................................................................... B-1
List of Tables....................................................................................... B-3
Appendix C
GNU Lesser General Public License.............................................................C-1
GNU Lesser General Public License ................................................... C-1
43iQ Instruction Manual
vii
Chapter 1
Introduction
The Thermo Scientific™ 43iQ Sulfur Dioxide (SO2) utilizes pulsed
fluorescence technology to measure the amount of sulphur dioxide in the
air.
The pulsing of the UV source lamp serves to increase the optical intensity
whereby a greater UV energy throughput and lower detectable SO2
concentration are realized.
Reflective bandpass filters, as compared to commonly used transmission
filters, are less subject to photochemical degradation and more selective in
wavelength isolation. This results in both increased detection specificity
and long term stability.
iQ Series
Instrument
Platform
Thermo Scientific
The iQ Series Instrument Platform is a smart environmental monitoring
solution for ambient and source gas analysis that affords greater control
over instrument performance and data availability.
●
Distributed Measurement and Control (DMC) module design
simplifies serviceability. Each DMC module contains its own
microprocessor control enabling functional performance validation
at the module level.
●
Built-in predictive diagnostics and preventive maintenance
schedules identify problems before they occur. The iQ Series
platform sends email notifications directly to Thermo Fisher
Scientific’s world class service support team or locally identified
addressees in order to proactively communicate analyzer
performance conditions and identify spare parts needs before an
operational concern arises.
●
The iQ Series platform supports Modbus, streaming and VNC
protocols over serial and Ethernet as well as analog and digital I/O
for easy integration into most data management systems.
●
Three standard USB ports afford convenient data download
capability as well as the ability to connect additional hardware, such
as a computer keyboard or mouse.
●
The iQ Series GUI runs on a 7” color touchscreen display. The
GUI is highly flexible and can be customized to enable a tailored
43iQ Instruction Manual 1-1
Introduction
iQ Series Instrument Platform
experience to simplify daily operations. Custom designed ePort
software allows remote access to the analyzer with a PC. The ePort
control mirrors the same GUI look and feel as the instrument
touchscreen providing a speedy and familiar operational experience.
Figure 1–1. 43iQ Front
1-2
43iQ Instruction Manual
Thermo Scientific
Introduction
43iQ Principle of Operation
43iQ Principle of
Operation
The 43iQ operates on the principle that SO2 molecules absorb ultraviolet
(UV) light and become excited at one wavelength, then decay to a lower
energy state emitting UV light at a different wavelength. Specifically,
SO 2 h1 SO 2 * SO 2 h 2
The sample is drawn into the 43iQ through the SAMPLE bulkhead, as
shown in Figure 1–2. The sample flows through a hydrocarbon “kicker,”
which removes hydrocarbons from the sample by forcing the hydrocarbon
molecules to permeate through the tube wall. The SO2 molecules pass
through the hydrocarbon “kicker” unaffected.
The sample then flows into the fluorescence chamber, where pulsating UV
light excites the SO2 molecules. The condensing lens focuses the pulsating
UV light into the mirror assembly. The mirror assembly contains four
selective mirrors that reflect only the wavelengths which excite SO2
molecules.
As the excited SO2 molecules decay to lower energy states they emit UV
light that is proportional to the SO2 concentration. The bandpass filter
allows only the wavelengths emitted by the excited SO2 molecules to reach
the photomultiplier tube (PMT). The PMT detects the UV light emission
from the decaying SO2 molecules. The photodetector, located at the back
of the fluorescence chamber, continuously monitors the pulsating UV light
source and is connected to a circuit that compensates for fluctuations in the
UV light.
As the sample leaves the optical chamber, it passes through a flow sensor, a
capillary, and the “shell” side of the hydrocarbon kicker. The sample then
flows to the pump and is exhausted out the EXHAUST bulkhead of the
analyzer. The 43iQ outputs the SO2 concentration to the front panel
display and the analog outputs, and also makes the data available over the
serial or Ethernet connection.
Thermo Scientific
43iQ Instruction Manual
1-3
Introduction
43iQ Principle of Operation
Figure 1–2. 43iQ Flow Schematic
Figure 1–3. 43iQ Flow Schematic with Zero Span
1-4
43iQ Instruction Manual
Thermo Scientific
Introduction
Specifications
Specifications
Table 1–1 lists the specifications for the 43iQ.
Table 1–1. 43iQ Specifications
Thermo Scientific
Range
010 ppm
025 mg/m3
Extended Ranges
0100 ppm
0250 mg/m3
Zero Noise
1.0 ppb RMS (10 second averaging time)
0.5 ppb RMS (60 second averaging time)
0.25 ppb RMS (300 second averaging time)
Detection Limit
2 ppb (10 second averaging time)
1 ppb (60 second averaging time)
0.25 ppb (300 second averaging time)
Zero Drift
<0.5 ppb (24 hour)
Span Drift
±0.5% full-scale (24 hour)
Response Time
60 sec (10 second averaging time)
110 sec (60 second averaging time)
320 sec (300 second averaging time)
Linearity
±1% full-scale
Flow Rate
0.5 lpm (standard)
Interferences (EPA levels)
Less than lower detectable limit except for the following:
NO < 3 ppb, M-Xylene < 1ppb, H2O < 3% of reading
Operating Temperature Range
0–45 °C
Power Requirements
100–240 VAC 50/60 Hz
275 watts
Physical Dimensions
24 in (D) x 16.75 in (W) x 8.72 in (H) [609 mm (D) 425.45 mm
(W) x 221.48 mm (H)]
Weight
37.1 lbs
Analog I/O
4 Isolated Voltage Inputs 0–10 V
6 Isolated Analog Voltages Outputs, with 4 selectable
ranges
6 Isolated Analog Current Outputs, with 2 selectable ranges
Digital I/O
16 Digital Inputs (TTL)
8 Solenoid Driver Outputs
10 Digital Reed Relay Contact Outputs
Serial Ports
1 RS-232/485 port
43iQ Instruction Manual
1-5
Introduction
Specifications
1 RS-485 External Accessory port
Other Ports
3 Full Speed USB ports (one in front, two in rear)
1 Gigabit Ethernet port
Communication Protocols
MODBUS, Streaming
Approvals and Certifications
CE, TUV-SUD Safety, EPA
Table 1–2. 43iQ Optional Permeation Oven Specifications
1-6
43iQ Instruction Manual
Temperature Control
Three user selectable set points: 30, 35, 45 C
Temperature Stability
± 0.1 C
Warm-up Time
1 hour (permeation device can take 24 to 48 hours to stabilize)
Carrier Gas Flow
700 scc/min
Chamber size
Accepts permeation tubes up to 9 cm in total length; 1 cm in
diameter
Physical Dimensions
Contained inside the 43iQ
Power Requirements
24 VDC, 50 watts (in addition to the standard 43iQ)
Weight
Approximately five pounds (in addition to standard 43iQ)
Thermo Scientific
Introduction
Dimensions
Dimensions
Figure 1–4. Bench Mount Assembly (dimensions in inches [mm])
Thermo Scientific
43iQ Instruction Manual
1-7
Introduction
Dimensions
Figure 1–5. Rack Mount Assembly (dimensions in inches [mm])
1-8
43iQ Instruction Manual
Thermo Scientific
Introduction
Dimensions
Figure 1–6. Rack Mount Requirements
Figure 1–7. Rack Requirements Part 2
Thermo Scientific
43iQ Instruction Manual
1-9
Chapter 2
Installation and Setup
Installation and Setup describes how to unpack, setup, and start-up the
instrument. The installation should always be followed by instrument
calibration as described in the “Calibration” chapter of this manual.
Equipment Damage Do not attempt to lift the instrument by the cover or
other external fittings. ▲
Unpacking and
Inspection
The 43iQ is shipped complete in one container. If there is obvious damage
to the shipping container when you receive the instrument, notify the
carrier immediately and hold for inspection. The carrier is responsible for
any damage incurred during shipment.
Use the following procedure to unpack and inspect the instrument.
1. Remove the instrument from the shipping container and set it on a
table or bench that allows easy access to both the front and rear.
2. Remove the cover to expose the internal components. (See “Figure 2–
1” on page 2-2.)
3. Check for possible damage during shipment.
4. Check that all connectors and circuit boards are firmly attached.
5. Re-install the cover.
6. Remove any protective plastic material from the case exterior.
Thermo Scientific
43iQ Instruction Manual 2-1
Installation and Setup
Cover Removing and Replacing
Cover Removing
and Replacing
Use the following procedure to remove and replace the cover.
Equipment required:
Phillips screwdriver, #2
1. Unfasten the four 8-32 screws securing the cover (shipping screws).
2. Press in both latches located on top cover and hold while pulling up to
remove. Set upright.
Figure 2–1. Removing the Cover
3. To replace, align cover and drop in. Latches will automatically snap in
place.
2-2
43iQ Instruction Manual
Thermo Scientific
Installation and Setup
Mounting Options
Mounting
Options
Bench Mount
The instrument can be installed in the following configurations:
●
Bench Mount
●
Rack Mount
Positioned on bench, includes installing feet. See Figure 2–2.
Equipment required:
Slot drive, 5/16-inch
1. Fasten feet in position 1 or 2 to fit to the desired depth.
Figure 2–2. Installing Feet
Thermo Scientific
43iQ Instruction Manual
2-3
Installation and Setup
Mounting Options
Rack Mount
Mounting in a rack includes removing the front panel and installing ears
and handles.
Equipment required:
Phillips drive, #2
1. Start by gripping from the top corners of the front panel and pull
outwards.
Figure 2–3. Removing the Front Panel
2. Unfasten the four 8-32 pan head screws.
3. Slide ears outwards.
4. Use the same four 8-32 pan head screws to secure it.
5. Install the handles with the four 8-32 flat head screws that came with
the handle kit on the backside as shown.
2-4
43iQ Instruction Manual
Thermo Scientific
Installation and Setup
Mounting Options
Figure 2–4. Installing Ears and Handles
Thermo Scientific
43iQ Instruction Manual
2-5
Installation and Setup
Setup Procedure
Setup Procedure
Use the following procedure to setup the instrument:
1. Connect the sample line to the SAMPLE bulkhead on the rear panel
(Figure 2–5). Ensure that the sample line is not contaminated by dirty,
wet, or incompatible materials. All tubing should be constructed of
PTFE, 316 stainless steel, borosilicate glass, or similar tubing with an
OD of 1/4-inch and a minimum ID of 1/8-inch. The length of the
tubing should be less than 10 feet.
Note Gas must be delivered to the instrument free of particulates. It may
be necessary to use the PTFE particulate filter as described in “PTFE
Particulate Filter” on page 9-23. ▲
Note If the sample may contain particulates larger than 5 microns, it
should be filtered before introducing it to the instrument. Use a filter (such
as PTFE) that does not interact with SO2 in the sample. If a sample filter is
used, all calibrations and span checks must be performed through the filter.
The filter element should be replaced regularly to prevent the absorption of
SO2 by trapped material on the filter. ▲
Note Gas must be delivered to the instrument at atmospheric pressure. It
may be necessary to use an atmospheric bypass plumbing arrangement as
shown in Figure 2–6 if gas pressure is greater than atmospheric pressure. ▲
2. Connect the EXHAUST bulkhead to a suitable vent. The exhaust line
should be 1/4-inch OD with a minimum ID of 1/8-inch. The length of
the exhaust line should be less than 10 feet. Verify that there is no
restriction in this line.
3. If the optional zero/span solenoid valves are installed, connect a source
of SO2 and HC free air to the ZERO IN bulkhead, and connect a
source of SO2 span gas to the SPAN bulkhead.
4. Connect a suitable recording device to the rear panel connector. For
detailed information about connecting to the instrument, refer to:
“Connecting External Devices” on page 9-1
Communications > “Analog I/O” on page 3-80, and “Digital I/O”
on page 3-81.
2-6
43iQ Instruction Manual
Thermo Scientific
Installation and Setup
Setup Procedure
5. Plug the instrument into an outlet of the appropriate voltage and
frequency.
Note If instrument is equipped with an internal permeation oven, refer to
Chapter 9, “Optional Equipment” for setup instructions. ▲
Icon Here
The 43iQ is supplied with a three-wire grounding cord. Under no
circumstances should this grounding system be defeated. ▲
Figure 2–5. 43iQ Rear Panel
Figure 2–6. Atmospheric Dump Bypass Plumbing
Thermo Scientific
43iQ Instruction Manual
2-7
Installation and Setup
Startup
Startup
Use the following procedure when starting the instrument.
1. Turn the power ON.
2. Allow 90 minutes for the instrument to stabilize.
3. Set instrument parameters such as operating ranges and averaging times
to appropriate settings. For more information about instrument
parameters, see the “Operation” chapter.
4. Before beginning the actual monitoring, perform a calibration as
described in the “Calibration” chapter.
Figure 2–7. Front Panel and Touchscreen Display
2-8
43iQ Instruction Manual
Thermo Scientific
Chapter 3
Operation
This chapter describes the functionality of the touchscreen user interface.
Instrument
Display
The Instrument Display consists of a Title Bar, a User Interface, and a
Status Bar. The Title Bar, located at the top, includes the Home button,
instrument name, instrument gas mode, and Help button. The User
Interface, located in the middle, is where the Home Screen and all other
screens are accessed. The Home Screen has three Main Menu buttons,
located on the left side, which include Calibration, Data, and Settings,
while the user interface to the right of the buttons displays the chemical
name, concentration value and unit. The Status Bar, located at the bottom,
includes the Back button, Access Levels, Health Check, Favorites, Date and
Time, and Contact Information.
Home Screen (single range mode)
Title Bar
User Interface
Status Bar
Thermo Scientific
43iQ Instruction Manual 3-1
Operation
Instrument Display
Home Screen (dual or auto range mode)
Title Bar
User Interface
Status Bar
3-2
43iQ Instruction Manual
Thermo Scientific
Operation
Instrument Display
The Instrument Display contains the following information:
●
●
●
Thermo Scientific
Title Bar:
●
Home button: When pressed, it brings you to the Home Screen.
●
Title Text: Displays instrument name when in the Home Screen.
Displays the chemical name, current concentration reading and
unit when in all other screens. When unit is pressed, it brings you
to the gas unit selection screen.
●
Gas Mode button: Displays current gas mode of the instrument.
When pressed, brings you to the Gas Mode selection screen.
●
Help button: When pressed, brings you to the help screens.
User Interface:
●
Calibration button: Allows the user to calibrate the instrument,
setup automatic calibrations, and view calibration data.
●
Data button: Allows the user to view, graph, stream, and analyze
data.
●
Settings button: Shows real-time status and alarms, also predictive
diagnostics and maintenance history. Contains controls for
operating the instrument, communications, and sets instrument
options.
●
Concentration: When in single mode, displays SO2 concentrations
in big, bold characters, depending on operating mode. When in
dual or auto mode, displays either high range or low range values
based on the range setting.
Status Bar:
●
Back button: When pressed, it displays the previous screen.
●
Access Levels button: Allows the user to set security access levels, and
allows/restricts access to functionality depending on the selected
access level.
●
Health Check button: Brings the user to the Health Check screen.
●
Favorites button: Allows user-selectable favorite buttons. To add to
the favorites screen, user presses the desired screen button for 2
seconds. The user will be directed to the favorites screen where the
user chooses the button position. To remove a favorite button from
the favorites screen, press and hold button for 2 seconds.
●
Clock: Displays current date and time.
43iQ Instruction Manual
3-3
Operation
Instrument Display
●
3-4
43iQ Instruction Manual
Thermo Scientific Information button: Shows contact information.
Thermo Scientific
Operation
Instrument Display
Main Menus and
Keypads
The Main Menu buttons, located on the Home Screen, contains three
submenus. Each submenu contains related instrument settings. This
chapter describes each submenu and screen in detail. Refer to the
appropriate sections for more information.
Calibration
Calibrate Background
Calibrate Span Coefficient
Zero/Span Schedule
Advanced Calibration
Manual Calibration
Adjust Background
Adjust Span Coefficient
Reset Bkg and Span Coef
Calibration History
Data
View Data Log (Last Hour)
Graph
View Data Log (Last 24 Hours)
Graph
View Data (User Defined Time)
Start Time
End Time – View Data
Graph
Advanced Data
Data Logging Setup
Streaming Data Setup
Settings
Health Check
Status and Alarms
Predictive Diagnostics
Maintenance
File Sharing and Support
Firmware Version
Measurement Settings
Averaging Time
Range Mode Selection
Range Setting
Gas Mode
Gas Units
Dilution Ratio
Advanced Measurement
Optical Bench Settings
Flash Lamp
Optical Span Test
Extended Ranges
Compensation
Perm Oven Settings
(Optional)
Pressure Calibration
Communication Settings
Instrument Settings
Display Setup
Alarm Setpoints
Language
Clock
Pump Power
Configuration
Security Access Levels
USB Drive
Firmware Update Via USB
Drive
Download Data to USB Drive
Change USB Password
User Contact Information
Update Bootloader
Thermo Scientific
43iQ Instruction Manual
3-5
Operation
Instrument Display
Numeric Keypad
3-6
43iQ Instruction Manual
User enters a value into the box using the number keypad. When the user
needs to change a value, such as for flow rates, temperatures or pressures,
the keypad screen will automatically display. Initially, the box above the
keypad will display the current value. Enter a new value using the keypad,
and then select the Enter button to set the new value or press the Cancel
button to exit the keypad screen and return to the previous screen without
saving the value.
Thermo Scientific
Operation
Instrument Display
Alphanumeric Keypad
Thermo Scientific
User enters a value into the box using the keypad. When the user needs to
change an alphanumeric value, this keypad will automatically display.
Initially, the box above the keypad will display the current value. Enter a
new value using the keypad, and then select the Enter button to set the
new value or press the Cancel button to exit the keypad screen and return
to the previous screen without saving the value. The alphanumeric keypad
is only available when the user needs to enter alphabet characters.
43iQ Instruction Manual
3-7
Operation
Instrument Display
3-8
43iQ Instruction Manual
Thermo Scientific
Operation
Calibration
Calibration
The Calibration screen allows the user to calibrate the system, setup
automatic calibrations, and view calibration data. See Chapter 4
“Calibration” for further instructions on how to run a calibration.
The following screens show the calibration screens in single range mode
and dual or auto range mode. The dual and auto range modes have two
SO2 span factors (high and low). This allows each range to be calibrated
separately. This is necessary if the two ranges used are not close to one
another. For example, a low SO2 range of .5 ppm and a high SO2 range of
10 ppm. For more information about range modes, see “Range Mode
Selection” on page 3-65.
Home Screen>Calibration (single range mode)
Home Screen>Calibration (dual or auto range mode)
The Calibration screen contains the following information:
Thermo Scientific
●
Calibrate Background: Sets the SO2 reading to zero.
●
Calibrate Span Coefficient: Sets the span coefficient when in single
range mode.
43iQ Instruction Manual
3-9
Operation
Calibration
3-10
43iQ Instruction Manual
●
Calibrate High Range Span Coefficient: Sets the high range span
coefficient when in dual or auto range mode.
●
Calibrate Low Range Span Coefficient: Sets the low range span
coefficient when in dual or auto range mode.
●
Zero/Span Schedule: Programs the instrument to perform fully
automated zero and span checks or adjustments.
●
Advanced Calibration: Calibrates the instrument using manual
zero/span calibration and provides calibration history.
Thermo Scientific
Operation
Calibration
Calibrate
Background
The Calibrate Background screen is used to calibrate the instrument zero
background. Before making an adjustment, be sure the analyzer samples
zero air for at least 5 minutes.
It is important to note the averaging time when calibrating. The longer the
averaging time the more precise the calibration results. To achieve
maximum precision, allow the instrument to stabilize each time input gas is
changed and set the averaging time to 300-second averaging.
Home Screen>Calibration>Calibrate Background
The Calibrate Background screen contains the following information:
Thermo Scientific
●
Target Concentration: Read only. Displays what the concentration value
will become when the calibrate button is pressed.
●
Current Background: Read only. Displays what the current user-set
background is.
●
Calculated Background: Read only. Displays what the current user-set
background will become when the calibrate button is pressed.
●
Calibrate: When pressed, updates the background value, setting the
displayed concentration to zero.
43iQ Instruction Manual
3-11
Operation
Calibration
Calibrate Span
Coefficient
The Calibrate Span Coefficient screen is used to enter the span
concentration and calibrate the SO2 span coefficient. The SO2 span
coefficient is calculated, stored, and used to correct the current reading.
The following screens are shown in single range mode and dual or auto
range mode. In dual or auto range modes, “High” or “Low” is displayed to
indicate the calibration of the high or low coefficient. The Calibrate High
Range Span Coefficient and Calibrate Low Range Span Coefficient screens
function the same way.
It is important to note the averaging time when calibrating. The longer the
averaging time the more precise the calibration results. To achieve
maximum precision, allow the instrument to stabilize each time input gas is
changed and set the averaging time to 300-second averaging.
Home Screen>Calibration>Calibrate Span Coefficient (single range
mode)
Home Screen>Calibration>Calibrate Span Coefficient (dual or auto
range mode)
3-12
43iQ Instruction Manual
Thermo Scientific
Operation
Calibration
The Calibrate Span Coefficient screen contains the following information:
Thermo Scientific
●
Edit Span Concentration: User enters the span gas concentration when
in single range mode.
●
Edit High Range Span Concentration: User enters the high range span
concentration when in dual or auto range mode.
●
Edit Low Range Span Concentration: User enters the low range span
concentration when in dual or auto range mode.
●
Current High Range Concentration: Read only. Current high range
concentration reading when in dual or auto range mode.
●
Current Low Range Concentration: Read only. Current low
concentration reading when in dual or auto range mode.
●
Current Span Coefficient: Read only. Displays the current instrument
span coefficient.
●
Current High Range Span Coefficient: Read only. Displays the current
instrument high range span coefficient.
●
Current Low Range Span Coefficient: Read only. Displays the current
instrument low range span coefficient.
●
Calculated Span Coefficient: Read only. After the “Edit Span
Concentration” value is entered, the new calculated span coefficient is
displayed.
●
Calculated High Range Span Coefficient: Read only. After the “Edit
High Range Span Concentration” value is entered, the new calculated
high range span coefficient is displayed.
●
Calculated Low Range Span Coefficient: Read only. After the “Edit Low
Range Span Concentration” value is entered, the new calculated low
range span coefficient is displayed.
●
Calibrate: When pressed, updates the coefficient and the concentration
should match the span concentration.
43iQ Instruction Manual
3-13
Operation
Calibration
Zero/Span Schedule
The Zero/Span Schedule is used to program the instrument to perform
fully automated zero and span checks or adjustments.
Home Screen>Calibration>Zero/Span Schedule
Home Screen>Calibration>Zero/Span Schedule>More
The Zero/Span Schedule contains the following information:
3-14
43iQ Instruction Manual
●
Zero/Span Schedule: Toggles zero/span schedule Enabled or Disabled.
●
Next Time: Allows the user to view and set the initial date and time
(24-hour format) of the zero/span schedule.
●
Period: Defines the period or interval between zero/span checks or
calibrations. If period = 0, the schedule runs continuously.
●
Zero Duration: Sets how long zero air is sampled by the instrument.
●
Span Duration: Sets how long span gas is sampled by the instrument.
●
Purge Duration: Sets how long the purge period will be at the end of
the schedule.
Thermo Scientific
Operation
Calibration
Thermo Scientific
●
Total Duration: Read only. Displays the total time duration of all
scheduled events.
●
Schedule Averaging Time: Allows the user to adjust the zero/span
schedule averaging time. This averaging time only affects the zero/span
schedule.
●
Background Calibration: Toggles Enabled/Disabled. If enabled,
background value is calibrated. If disabled, schedule runs a background
check only and background value is not updated.
●
Span Calibration: Toggles Enabled/Disabled. If enabled, span
coefficient is calibrated. If disabled, schedule runs a calibration check
only and span coefficient is not updated.
●
Zero : Span Ratio: Allows the user to perform more scheduled
background calibration checks to span calibration checks. Default is 1
and therefore reads 1:1. (This means that each time the schedule is run,
both the zero duration and span duration occurs.) The zero/span ratio
is allowable between 1 to 99. If 99 is chosen, the schedule should only
perform the Span on the 99th iteration.
43iQ Instruction Manual
3-15
Operation
Calibration
Advanced
Calibration
The Advanced Calibration screen provides a manual way to calibrate the
instrument and view the calibration history. See Chapter 4 “Calibration”
for further instructions on how to run a calibration.
Home Screen>Calibration>Advanced Calibration
The Advanced screen contains the following information:
3-16
43iQ Instruction Manual
●
Manual Calibration: The user manually calibrates the background or
span coefficient.
●
Calibration History: Lists all calibrations performed and calibration
checks.
Thermo Scientific
Operation
Calibration
Manual Calibration
The Manual Calibration screen adjusts the zero background or span
coefficient based on a user entered value. See Chapter 4, “Calibration” for
instructions on how to run a Manual Calibration.
The following screens show the manual calibration screens in single range
mode and dual or auto range mode. In dual or auto range modes, “High
Range” or “Low Range” buttons are displayed to indicate the calibration of
the high or low coefficient.
Home Screen>Calibration>Advanced Calibration>Manual Calibration
(single range mode)
Home Screen>Calibration>Advanced Calibration>Manual Calibration
(dual or auto range mode)
The Manual Calibration screen contains the following information:
Thermo Scientific
●
Adjust Background: Allows the user to manually adjust the zero
background.
●
Adjust Span Coefficient: Allows the user to manually adjust the span
coefficient when in single range mode.
43iQ Instruction Manual
3-17
Operation
Calibration
3-18
43iQ Instruction Manual
●
Adjust High Range Span Coefficient: Allows the user to manually adjust
the high range span coefficient when in dual or auto range mode.
●
Adjust Low Range Span Coefficient: Allows the user to manually adjust
the low range span coefficient when in dual or auto range mode.
●
Reset Background to 0.000 and Span Coefficient to 1.000: Resets all
backgrounds and coefficients.
Thermo Scientific
Operation
Calibration
Adjust Background
The Adjust Background screen is used to manually adjust the zero
background.
Home Screen>Calibration>Advanced Calibration>Manual
Calibration>Adjust Background
The Adjust Background screen contains the following information:
Thermo Scientific
●
Adjust Background: User manually adjusts zero background.
●
Adjusted Concentration: Read only. Shows adjusted concentration based
on adjusted zero background.
●
Calibrate: Calibrates zero background by saving the newly adjusted zero
background value.
43iQ Instruction Manual
3-19
Operation
Calibration
Adjust Span Coefficient
The Adjust Span Coefficient screen is used to manually adjust the span
coefficient.
The following screen is shown in single range mode and dual or auto range
mode. In dual or auto range modes, “High Range” or “Low Range” is
displayed to indicate the calibration of the high or low coefficient. The
Adjust High Range Span Coefficient and Adjust Low Range Span
Coefficient screens function the same way.
Home Screen>Calibration>Advanced>Manual Calibration>Adjust
Span Coefficient (single range mode)
Home Screen>Calibration>Advanced Calibration>Manual
Calibration>Adjust High Range Span Coefficient (dual or auto range
mode)
The Adjust Span Coefficient screen contains the following information:
3-20
43iQ Instruction Manual
●
Adjust Span Coefficient: User manually adjusts span coefficient when in
single range mode.
●
Adjusted Concentration: Read only. Shows adjusted concentration based
on adjusted span coefficient when in single range mode.
Thermo Scientific
Operation
Calibration
Thermo Scientific
●
Adjust High Range Span Coefficient: User manually adjusts high range
span coefficient when in dual or auto range mode.
●
Adjusted High Range Concentration: Read only. Shows adjusted high
range concentration based on adjusted high range span coefficient
when in dual or auto range mode.
●
Adjust Low Range Span Coefficient: User manually adjusts low range
span coefficient when in dual or auto range mode.
●
Adjusted High Range Concentration: Read only. Shows adjusted low
range concentration based on adjusted low range span coefficient when
in dual or auto range mode.
●
Calibrate: Calibrates span coefficient by saving the newly adjusted span
coefficient.
43iQ Instruction Manual
3-21
Operation
Calibration
Calibration History
The Calibration History screen shows the log of calibrations and
calibration checks performed.
Use the and buttons to move up and down and the and
buttons to move left and right.
Home Screen>Calibration>Advanced Calibration>Calibration History
Note Pressing the Calibration History button responds with Retrieving
calibration log data, it may take a few seconds... ▲
The Calibration History screen contains the following information:
3-22
43iQ Instruction Manual
●
Time Stamp: Time of calibration or calibration check.
●
Event: Lists the type of calibration event.
●
Result: Concentration result.
●
Target: Concentration setpoint value.
●
Units: Displays units for each item.
●
Average Time: Averaging time used during the calibration or calibration
check.
Thermo Scientific
Operation
Data
Data
The Data screen is used to view recorded concentrations and instrument
data. Users can view both tabular data and graphed data.
Home Screen>Data
The Data screen contains the following information:
Thermo Scientific
●
View Data Log (Last Hour): User views last hour of historical data.
Table shows most recent data on top.
●
View Data Log (Last 24 Hours): User views 24-hour of historical data.
Table shows most recent data on top.
●
View Data Log (User Defined Time): User selects the start and end time
for viewing the data. Table shows most recent data on top.
●
Advanced Data Setup: Allows the user to set up the parameters of how
the data is stored.
43iQ Instruction Manual
3-23
Operation
Data
View Data Log (Last
Hour)
The View Data Log (Last Hour) screen allows the user to view the last hour
worth of data in tabular and/or graphical form.
Use the and buttons to move up and down and the and
buttons to move left and right.
Home Screen>Data>View Data Log (Last Hour)
Note Pressing the View Data Log (Last Hour) responds with Retrieving
user log data, it may take a few seconds... ▲
The View Data Log (Last Hour) screen contains the following options:
●
3-24
43iQ Instruction Manual
Graph: Displays data graph for the column selected. The graph time
axis is defined by the data set in the table.
Thermo Scientific
Operation
Data
View Data Log (Last
24 Hours)
The View Data Log (Last 24 Hours) screen allows the user to instantly
view the last 24 hours worth of data in real time.
Use the and buttons to move up and down and the and
buttons to move left and right.
Home Screen>Data>View Data Log (Last 24 Hours)
Note Pressing the View Data Log (Last 24 Hours) responds with
Retrieving user log data, it may take a few seconds... ▲
The View Data Log (Last 24 Hours) screen contains the following options:
●
Thermo Scientific
Graph: Displays data graph for the column selected. The graph time
axis is defined by the data set in the table.
43iQ Instruction Manual
3-25
Operation
Data
View Data Log (User
Defined Time)
The View Data (User Defined Time) screen is used to specify the start and
end time for viewing the data logging table.
Home Screen>Data>View Data Log (User Defined Time)
Home Screen>Data>View Data Log (User Defined Time)>Save Data
Logging Start Time
The View Data Log (User Defined Time) screen contains the following
information:
3-26
43iQ Instruction Manual
●
Month: Sets month of data logging start time.
●
Day: Sets day of data logging start time.
●
Year: Sets year of data logging start time.
●
Hours: Sets hours of data logging start time.
●
Minutes: Sets minutes of data logging start time.
●
Save Data Logging Start Time: Pressing this button saves the start time
and follows directly to the end time selection for the data logging
screen.
Thermo Scientific
Operation
Data
The View Data Log (User Defined Time) End Time screen contains the
following information:
●
Month: Sets month of data logging end time.
●
Day: Sets day of data logging end time.
●
Year: Sets year of data logging end time.
●
Hours: Sets hour of data logging end time.
●
Minutes: Sets minute of data logging end time.
●
Save Data Logging End Time: Pressing the Save Data Logging End
Time button saves the end time and follows directly to the data logging
table.
Note End time should not be greater than 1 year from start time . ▲
Note The datalogging table is limited to 10,000 points. ▲
Thermo Scientific
43iQ Instruction Manual
3-27
Operation
Data
Advanced Data
Setup
The Advanced Data Setup screen allows the user to select variables and set
up parameters for data logging and streaming data.
Home Screen>Data>Advanced Data Setup
The Advanced Data Setup screen contains the following information:
3-28
43iQ Instruction Manual
●
Data Logging Setup: User selects the parameters for collecting logged
data.
●
Streaming Data Setup: User selects the parameters for streaming data to
a computer in real time.
Thermo Scientific
Operation
Data
Data Logging Setup
The Data Logging Setup screen allows the user to select data to be stored
and how it is stored.
Home Screen>Data>Advanced>Data Logging Setup
The Data Logging Setup screen contains the following information:
Thermo Scientific
●
Select Data Logging Variables: User selects instrument variables to log.
See “Table 3–1” for data logging variables list.
●
Period: User selects how often data is collected by setting the duration
between logged data values.
●
Data Treatment: Toggles between Average, Current, Minimum and
Maximum. When set to average, the average value during the period
will be recorded. When set to current, the latest data will be recorded.
When set to minimum or maximum, the minimum or maximum value
during the period will be recorded.
●
Erase Data Log Records: Allows the user to erase all values in the data log
and updates the data logging table.
43iQ Instruction Manual
3-29
Operation
Data
Select Data Logging
Variables
The Select Data Logging Variables screen allows the user to select which
variables to store. Note: The Data logging and Streaming variable lists are
exclusive from each other but contain the same variable selections.
Use the and buttons to scroll through the variables. Select the
variables to log by pressing the corresponding cells. Next, press the
Commit Changes button to save selections. Yellow buttons indicate that
the variable is selected. More than one can be chosen.
Home Screen>Data>Advanced>Data Logging Setup>Select Data
Logging Variables
The following table contains the variables that can be selected for data
logging:
Table 3–1. Data Logging Variables
Description
Concentration (ppb or μg/m3)
High Concentration (ppb or μg/m3)
Bench Pressure (mmHg)
Instrument Temperature (degC)
Flow (L/min)
Pump Pressure (mmHg)
Averaging Time (sec)
High Averaging Time (sec)
Concentration Alarm
Bench Pressure Alarm
Flow Alarm
Temperature Alarm
3-30
43iQ Instruction Manual
Thermo Scientific
Operation
Data
Auto Zero Cal/Check Alarm
Auto Span Cal/Check Alarm
Flow Pressure Module Alarms
Concentration Alarms
Gas Mode
General Alarm
Alerts
Instrument Error
Low Dynamic Filter Status
High Dynamic Filter Status
Dilution Ratio
Bench Alarm Count
Bench Temperature Alarm
Bench Temperature (degC)
Flasher High Voltage (Volts)
PMT High Voltage (Volts)
External Alarm 1
External Alarm 2
External Alarm 3
Analog Input 1
Analog Input 2
Analog Input 3
Analog Input 4
Analog Alarms
PSB Alarms
Perm Oven Body Temp (deg C)
Perm Oven Gas Temp (deg C)
Thermo Scientific
43iQ Instruction Manual
3-31
Operation
Data
Streaming Data Setup
The Streaming Data Setup screen allows the user to stream data to a
computer.
Home Screen>Data>Advanced>Streaming Data Setup
The Streaming Data Setup screen contains the following information:
3-32
43iQ Instruction Manual
●
Select Streaming Variables: User selects which variables to stream. See
Table 3–2 for streaming variable list.
●
Period: Sets the time between streamed data.
●
Show Labels: Toggles on/off. When on, shows variable labels to the left
of the variable values.
●
Show Timestamp: Toggles on/off. When on, shows timestamp at the
beginning of each row of data.
Thermo Scientific
Operation
Data
Select Streaming
Variables
The Select Streaming Variables screen allows the user to select which
variables to track. Note: The Data logging and Streaming variable lists are
exclusive from each other but contain the same variable selections.
Use the and buttons to scroll through the variables. Select the
variables to log by pressing the corresponding cells. Next, press the
Commit Changes button to save selections. Yellow buttons indicate that
the variable is selected. More than one can be chosen.
Home Screen>Data>Advanced>Streaming Data Setup>Select
Streaming Variables
The following table contains the variables that can be selected for streaming
data:
Table 3–2. Streaming Data Variables
Description
Concentration (ppb or μg/m3)
High Concentration (ppb or μg/m3)
Bench Pressure (mmHg)
Instrument Temperature (degC)
Flow (L/min)
Pump Pressure (mmHg)
Averaging Time (sec)
High Averaging Time (sec)
Concentration Alarm
Bench Pressure Alarm
Flow Alarm
Thermo Scientific
43iQ Instruction Manual
3-33
Operation
Data
Description
Temperature Alarm
Auto Zero Cal/Check Alarm
Auto Span Cal/Check Alarm
Flow Pressure Module Alarms
Concentration Alarms
Gas Mode
General Alarm
Alerts
Instrument Error
Low Dynamic Filter Status
High Dynamic Filter Status
Dilution Ratio
Bench Alarm Count
Bench Temperature Alarm
Bench Temperature (degC)
Flasher High Voltage (Volts)
PMT High Voltage (Volts)
External Alarm 1
External Alarm 2
External Alarm 3
Analog Input 1
Analog Input 2
Analog Input 3
Analog Input 4
Analog Alarms
PSB Alarms
Perm Oven Body Temp (deg C)
Perm Oven Gas Temp (deg C)
3-34
43iQ Instruction Manual
Thermo Scientific
Operation
Settings
Settings
The Settings screen allows the user to view the status and alarms, set up
user preferences, communicate with outside devices and computers,
download files to USB, and sets security protocol.
Home Screen>Settings
The Settings screen contains the following information:
Thermo Scientific
●
Health Check: View instrument status and alarms, predictive
diagnostics, preventive maintenance alerts, maintenance history, email
health check report files, and contact Thermo Fisher Scientific
technical support.
●
Measurement Settings: Allows the user to setup user preferences as
related to the concentration readings.
●
Communications: Allows the user to communicate with outside devices.
●
Instrument Setting: Allows the user to setup alarm setpoints and user
preferences.
●
Configuration: User selects which options to enable.
●
Security Access Levels: User selects security protocol. User can also
change security passwords.
●
USB Drive: User can update instrument firmware, download data, and
change USB password.
●
User Contact Information: User sets up their contact information.
●
Update Bootloader: Used to update bootloader when an update to the
bootloader is available.
43iQ Instruction Manual
3-35
Operation
Settings
Health Check
The Health Check screen is used for viewing instrument status and alarms,
predictive diagnostics, preventive maintenance schedules, maintenance
history, emailing files describing the health/status of the instrument, and
viewing the instrument’s firmware version.
Note
This symbol denotes there is an active alarm in the module. ▲
Note
This symbol denotes there is an active maintenance alarm or
condition in the module. ▲
Home Screen>Settings>Health Check
The Health Check screen contains the following information:
3-36
43iQ Instruction Manual
●
Status and Alarms: Allows the user to view the status and alarm menus.
Menus are broken down according to modules where the user can view
instrument readings, setpoints and alarms.
●
Predictive Diagnostics: Smart module diagnostics, which shows possible
future issues.
●
Maintenance History: Allows the user to set up a maintenance schedule
and track maintenance history.
●
File Sharing and Support: File sharing via email. Support through
Thermo Fisher Scientific technical support.
●
Firmware Version: Shows the instrument’s firmware version.
Thermo Scientific
Operation
Settings
Status and Alarms
The Status and Alarms screen provides information with respect to module
alarms. In each screen, instrument readings, setpoints, and low/high alarm
values are displayed. If applicable, setpoints and alarms are also settable
from the Settings>Instrument Settings screen.
Note
This symbol denotes there is an active alarm in the module. ▲
Home Screen>Settings>Health Check>Status and Alarms
The Status and Alarms screen contains the following information:
Thermo Scientific
●
Concentration: Displays SO2 concentrations, alarms, and calibration
pass/fail status.
●
SO2 Bench: Displays bench module alarms and faults.
●
Perm Oven: Displays perm oven module alarms and faults.
●
Flow and Pressure: Displays flow and pressure alarms and faults.
●
Peripherals Support: Displays peripherals support alarms and faults.
●
Analog I/O: Displays analog input/output alarms and faults.
●
Digital I/O: Displays digital input/output alarms and faults.
●
Valve and Pump Resets: User can reset valve and pump power.
●
Serial Numbers: Displays all the serial numbers for the instrument.
43iQ Instruction Manual
3-37
Operation
Settings
Concentration
The Concentration screen provides status and alarms for SO2
concentration, background cal/checks, and span cal/checks. If the item
being monitored goes outside the lower or higher alarm limit, an alarm is
activated.
Use the and buttons to move up and down and the and
buttons to move left and right.
Home Screen>Settings>Health Check>Status and
Alarms>Concentration
The Concentration screen contains the following information:
●
●
3-38
43iQ Instruction Manual
Across:
●
Concentration: This column lists items associated with the SO2
concentration.
●
Value: Displays the current value for each item.
●
Low Alarm: Displays low alarm for each item.
●
High Alarm: Displays high alarm for each item.
●
Span Conc: Span concentration used in the span calibration or span
check.
●
Units: Displays units for each item.
Down:
●
SO2: SO2 concentration.
●
Bkg Check Offset: Displays concentration based on the last
attempted background calibration. High alarm shows user defined
limit for acceptable background check offset.
Thermo Scientific
Operation
Settings
●
Span Check Offset: Displays concentration based on the last
attempted span calibration. High alarm shows user defined limit for
acceptable span check offset (compared to the span concentration
value). Span concentration shows span setpoint.
Note If both the low alarm and high alarms are set to zero, then no alarm
will show. ▲
Thermo Scientific
43iQ Instruction Manual
3-39
Operation
Settings
SO2 Bench
The SO2 Bench screen provides status and alarms related to the SO2 bench
module. If an item being monitored goes outside the lower or higher alarm
limit, an alarm is activated.
Use the and buttons to move up and down and the and
buttons to move left and right.
Home Screen>Settings>Health Check>Status and Alarms>SO2 Bench
The SO2 Bench screen contains the following information:
●
●
3-40
43iQ Instruction Manual
Across:
●
SO2 Bench: This column lists items associated with the SO2 bench.
●
Value: Displays the current value for each item.
●
Low Alarm: Displays low alarm for each item.
●
High Alarm: Displays high alarm for each item.
●
Units: Displays units for each item.
Down:
●
Chamber Temperature: Displays the current chamber temperature.
User can adjust low and high alarm limits.
●
Lamp Intensity: Displays the current lamp intensity reading. User
can adjust low and high alarm limits.
●
Lamp Voltage: Displays the current lamp voltage. User can adjust
low and high alarm limits.
●
PMT Supply Voltage: Displays the current PMT supply voltage.
●
Bench Pressure: Displays the current bench pressure reading.
●
Flow: Displays the current sample flow reading.
Thermo Scientific
Operation
Settings
Thermo Scientific
●
Instrument Temp: Displays the current instrument temperature
reading.
●
Board Communication: Displays OK/Fail for board communication
status.
●
Power Supply: Displays OK/Fail of power supplies. Power supply
goes red if any voltages are outside their limits. No voltage rows
ever get highlighted.
●
3.3V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
●
5V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
●
15V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
●
24V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
43iQ Instruction Manual
3-41
Operation
Settings
Perm Oven
The Perm Oven screen provides status and alarms for the permeation oven,
if installed. If an item being monitored goes outside the lower or higher
alarm limit, an alarm is activated.
Use the and buttons to move up and down and the and
buttons to move left and right.
Home Screen>Settings>Health Check>Status and Alarms>Perm Oven
The Perm Oven screen contains the following information:
●
●
3-42
43iQ Instruction Manual
Across:
●
Perm Oven: This column lists items associated with the perm oven.
●
Value: Displays the current value for each item.
●
Low Alarm: Displays low alarm for each item.
●
High Alarm: Displays high alarm for each item.
●
Units: Displays units for each item.
Down:
●
Perm Oven Temperature: Displays the current perm oven
temperature. User can adjust low and high alarm limits.
●
Board Communication: Displays OK/Fail for board communication
status.
●
Power Supply: Displays OK/Fail of power supplies. Power supply
goes red if any voltages are outside their limits. No voltage rows
ever get highlighted.
●
Heater Power Diagnostics: Displays current heater voltage readings.
Alarm limits are not changeable.
Thermo Scientific
Operation
Settings
Thermo Scientific
●
5 V Diagnostics: Displays current voltage readings. Alarm limits are not
changeable.
●
3.3 V Diagnostics: Displays current voltage readings. Alarm limits are
not changeable.
●
3 V Diagnostics: Displays current voltage readings. Alarm limits are not
changeable.
●
2.5 V Diagnostics: Displays current voltage readings. Alarm limits are
not changeable.
43iQ Instruction Manual
3-43
Operation
Settings
Flow and Pressure
The Flow and Pressure screen provides status and alarms related to the flow
and pressure module. If an item being monitored goes outside the lower or
higher alarm limit, an alarm is activated.
Use the and buttons to move up and down and the and
buttons to move left and right.
Home Screen>Settings>Health Check>Status and Alarms>Flow and
Pressure
The Flow and Pressure screen contains the following information:
●
●
3-44
43iQ Instruction Manual
Across:
●
Flow and Pressure: This column lists items associated with the flow
and pressure module.
●
Value: Displays the current value for each item.
●
Low Alarm: Displays low alarm status for each item.
●
High Alarm: Displays high alarm status for each item.
●
Units: Displays units for each item.
Down:
●
Flow: Displays the current sample flow reading. User can adjust low
and high alarm limits.
●
Bench Pressure: Displays the current bench pressure reading. User
can adjust low and high alarm limits.
●
Pump Pressure: Displays the current pump pressure reading.
●
Instrument Temperature: Displays the current instrument
temperature reading. User can adjust low and high alarm limits.
Thermo Scientific
Operation
Settings
Thermo Scientific
●
Board Communication: Displays OK/Fail for board communication
status.
●
Power Supply: Displays OK/Fail of power supplies. Power supply
goes red if any voltages are outside their limits. No voltage rows
ever get highlighted.
●
2.5V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
●
3.3V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
●
5.0V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
●
24V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
43iQ Instruction Manual
3-45
Operation
Settings
Peripherals Support
The Peripherals Support screen provides status and alarms related to the
peripheral module. If an item being monitored goes outside the lower or
higher alarm limit, an alarm is activated.
Use the and buttons to move up and down and the and
buttons to move left and right.
Home Screen>Settings>Health Check>Status and Alarms> Peripherals
Support
The Peripherals Support screen contains the following information:
●
●
3-46
43iQ Instruction Manual
Across:
●
Peripherals Support: This column lists items associated with the
peripherals support module.
●
Value: Displays the current value for each item.
●
Low Alarm: Displays low alarm status for each item.
●
High Alarm: Displays high alarm status for each item.
●
Units: Displays units for each item.
Down:
●
Module Temperature: Displays the current module temperature of
the module.
●
Sample Valve: Displays whether or not the sample valve is activated.
●
Zero Valve: Displays whether or not the zero valve is activated.
●
Span Valve: Displays whether or not the span valve is activated.
●
Instrument Error: Displays OK/Fail for PCP, datalogging,
streaming, serial server, and Modbus protocols.
Thermo Scientific
Operation
Settings
Thermo Scientific
●
Board Communication: Displays OK/Fail for board communication
status.
●
Power Supply: Displays OK/Fail of power supplies. Power supply
goes red if any voltages are outside their limits. No voltage rows
ever get highlighted.
●
2.5V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
●
3.3V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
●
5.0V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
●
24V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
●
5.0V Step Board 1: Displays current voltage readings. Alarm
limits are not changeable.
●
24V Step Board 1: Displays current voltage readings. Alarm
limits are not changeable.
●
5.0V Step Board 2: Displays current voltage readings. Alarm
limits are not changeable.
●
24V Step Board 2: Displays current voltage readings. Alarm
limits are not changeable.
43iQ Instruction Manual
3-47
Operation
Settings
Valve and Pump Resets
The Valve and Pump Resets screen allows the user to reset a valve or pump
after a failure due to excessive amperage.
Note
This symbol denotes that the device needs to be reset. ▲
Note Resetting one valve will reset all valves. ▲
Home Screen>Settings>Health Check>Status and Alarms>Valve and
Pump Resets
The Valve and Pump Resets screen contains the following information:
3-48
43iQ Instruction Manual
●
Sample Valve Reset: Resets sample valve.
●
Zero Valve Reset: Resets zero valve.
●
Span Valve Reset: Resets span valve.
●
Pump Reset: Resets pump.
Thermo Scientific
Operation
Settings
Analog I/O
The Analog I/O screen provides status and alarms related to the analog
input/output module. If an item being monitored goes outside the lower or
higher alarm limit, an alarm is activated.
Use the and buttons to move up and down and the and
buttons to move left and right.
Home Screen>Settings>Health Check>Status and Alarms>Analog I/O
The Analog I/O screen contains the following information:
●
●
Thermo Scientific
Across:
●
Analog IO: This column lists items associated with the analog I/O.
●
Value: Displays the current value for each item.
●
Low Alarm: Displays low alarm status for each item.
●
High Alarm: Displays high alarm status for each item.
●
Units: Displays units for each item.
Down:
●
Voltage Output Channel 1–6: Displays real-time voltage output for
each channel.
●
Current Output Channel 1–6: Displays real-time current output for
each channel.
●
Chip Temperatures: Displays OK/Fail for chip temperatures.
●
Chip 1–3 Communication: Displays OK/Fail for each chip
communication.
●
Test Mode: Displays test mode on or off.
43iQ Instruction Manual
3-49
Operation
Settings
3-50
43iQ Instruction Manual
●
Board Communication: Displays OK/Fail for board communication
status.
●
Power Supply: Displays OK/Fail of power supplies. Power supply
goes red if any voltages are outside their limits. No voltage rows
ever get highlighted.
●
3.3V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
●
5.0V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
●
5.0V Ref Diagnostic: Displays current voltage readings. Alarm
limits are not changeable.
●
15V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
●
-15V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
Thermo Scientific
Operation
Settings
Digital I/O
The Digital I/O screen provides status and alarms related to the digital
input/output module. If an item being monitored goes outside the lower or
higher alarm limit, an alarm is activated.
Use the and buttons to move up and down and the and
buttons to move left and right.
Home Screen>Settings>Health Check>Status and Alarms>Digital I/O
The Digital I/O screen contains the following information:
●
●
Thermo Scientific
Across:
●
Digital IO: This column lists items associated with the digital I/O.
●
Value: Displays the current value for each item.
●
Reset: Resets item.
●
Low Alarm: Displays low alarm status for each item.
●
High Alarm: Displays high alarm status for each item.
●
Units: Displays units for each item.
Down:
●
Solenoid 1–8: Displays whether or not the solenoid is activated by
showing the current in mA.
●
External Alarm 1–3: Displays OK/Fail for external alarms.
●
Relay Test Mode: Displays relay test mode on or off.
●
Solenoid Test Mode: Displays solenoid test mode on or off.
●
Board Communication: Displays OK/Fail for communication status.
43iQ Instruction Manual
3-51
Operation
Settings
●
3-52
43iQ Instruction Manual
Power Supply: Displays OK/Fail of power supplies. Power supply
goes red if any voltages are outside their limits. No voltage rows
ever get highlighted.
●
3.3V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
●
5.0V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
●
24V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
Thermo Scientific
Operation
Settings
Serial Numbers
The Serial Numbers screen displays the serial number for each module.
Home Screen>Settings>Health Check>Status and Alarms>Serial
Numbers
The Serial Numbers screen contains the following information:
The Serial Numbers screen contains the following information:
Thermo Scientific
●
Instrument: Instrument serial number.
●
SO2 Bench: SO2 bench serial number.
●
Perm Oven: Optional perm oven serial number.
●
Flow and Pressure: Flow and pressure serial number.
●
Peripherals Support: Peripherals support serial number.
●
Analog I/O: Analog I/O serial number.
●
Digital I/O: Digital I/O serial number.
43iQ Instruction Manual
3-53
Operation
Settings
Predictive Diagnostics
The Predictive Diagnostics screen is a feature that allows the instruments to
anticipate maintenance needs, reduce downtime, and reduce
troubleshooting time. If button is greyed out, no maintenance is needed. If
button is blue, maintenance is suggested.
Note
This symbol denotes there is an active maintenance related
warning in the module. ▲
Home Screen>Settings>Health Check>Predictive Diagnostics
The Predictive Diagnostics screen contains the following information:
3-54
43iQ Instruction Manual
●
Sample Pump
●
Capillary
●
Flow Path
●
Sample Valve
●
Zero Valve
●
Span Valve
Thermo Scientific
Operation
Settings
Maintenance
The Maintenance screen reminds the user when certain instrument
components need to be serviced/replaced.
Home Screen>Settings>Health Check>Maintenance
The Maintenance screen contains the following information:
Thermo Scientific
●
Preventive Maintenance: Shows suggested service interval and time left
for component replacement.
●
Change Part: User logs component fix.
●
Maintenance History: Shows the log of all recorded component fixes.
●
Maintenance History: Shows the log of all recorded component fixes.
●
Advanced Maintenance: Resets all preventive maintenance items.
43iQ Instruction Manual
3-55
Operation
Settings
Preventive Maintenance
The Preventive Maintenance screen reminds the user when certain
instrument components need to be serviced/replaced. When the “Months
Left” has decreased to 1, the row is highlighted yellow. If the “Months
Left” is 0 or less, the row is highlighted red and the maintenance icon
(gears) will appear in the status bar located at the bottom of the screen.
Use the and buttons to move up and down and the and
buttons to move left and right.
Home Screen>Settings>Health Check>Maintenance>Preventive
Maintenance
The Preventive Maintenance screen contains the following information:
●
3-56
43iQ Instruction Manual
Across:
●
Component: Device to be routinely serviced or replaced.
●
Interval in Months: Expected period of time before a component
needs to be checked and/or serviced.
●
Months Left: Count down timer in months. Remaining time since
the beginning of the service interval. When the value is 1 or less,
the row will be highlighted and it is suggested that the component
should be checked and/or serviced.
●
Reset: Once the component is serviced/replaced, the user presses the
Reset button and the “Months Left” value resets to the “Service
Interval in Months” value.
●
Alert: Allows the user to opt out of receiving preventive
maintenance alerts. Displays Enabled/Disabled for each
component.
Thermo Scientific
Operation
Settings
●
Down:
●
Thermo Scientific
Optical Bench Module: Service interval for optical bench module
components.
●
UV Lamp: Service interval for UV lamp.
●
PMT: Service interval for PMT.
●
PMT Base Socket: Service interval for PMT base socket.
●
Flash Intensity Assembly: Service interval for flash intensity
assembly.
●
Mirrors: Service interval for mirrors.
●
Permeation Tube: Service interval for permeation tube.
●
Flow System: Service interval for the flow system components.
●
Capillaries: Service interval for capillaries.
●
Pump: Service interval for pump.
●
DC Power Supply: Service interval for DC power supply.
●
Fan Filter: Service interval for fan filter.
●
System Components: Service interval for system components.
●
Purafil: Service interval for purafil.
●
Charcoal: Service interval for charcoal.
●
Dri-Rite: Service interval for dri-rite.
43iQ Instruction Manual
3-57
Operation
Settings
Change Part
The Change Part screen allows the user to enter the component being
serviced and the type of fix. Pressing commit will update the preventive
maintenance table and predictive diagnostics screen when applicable.
Home Screen>Settings>Health Check>Maintenance>Change Part
The Change Part screen contains the following information:
3-58
43iQ Instruction Manual
●
Select Part: User selects part to service from the selection table.
●
Fix: User chooses from new, rebuilt, cleaned, and unknown.
●
Comment: User can write a brief comment, which will be saved to the
preventive maintenance history table.
●
Commit: User commits and saves the selected part fix.
Thermo Scientific
Operation
Settings
Maintenance History
The Maintenance History screen allows the user to view when components
are replaced, rebuilt, or cleaned. When a user changes a part in the change
part screen, the new row will be automatically created at the top in the
maintenance history table.
Use the and buttons to move up and down and the and
buttons to move left and right.
Home Screen>Settings>Health Check>Maintenance>Maintenance
History
Note Retrieving maintenance history data, it may take a few seconds... ▲
The Maintenance History screen contains the following information:
Thermo Scientific
●
Part: Component that has been fixed.
●
Fix: The type of maintenance.
●
Date: Shows date/time when service was logged.
●
Service Months: Amount of time in months since last service.
●
Comments: Shows comments entered from time of change.
43iQ Instruction Manual
3-59
Operation
Settings
File Sharing and Support
The File Sharing and Support screen allows the user to send health check
report files to Thermo Fisher Scientific technical support or user emails.
The Health Report file includes: Status and Alarms, PM Alerts, Activity
Log, Service Database, Cal History, and Data Log (last 24 hours).
Home Screen>Settings>Health Check>File Sharing and Support
The File Sharing and Support screen contains the following information:
3-60
43iQ Instruction Manual
●
Download Health Check Report to USB Drive: Sends the health report to
USB drive.
●
Email Health Check Report File to Technical Support: Sends the health
report file to technical support and the customer email addresses via
email.
●
Email Health Check Report to Personal Account: Sends the health report
file to a personal account via email.
●
iQ360: The iQ360 feature is a paid subscription enabling or disabling
the instrument to send automated emails to technical support when an
alarm or alert is triggered.
●
Request a Field Service Visit: Sends a field service visit to technical
support.
Thermo Scientific
Operation
Settings
Measurement
Settings
The Measurement Settings screen contains a number of submenus where
instrument parameters and settings can be read and modified.
The following screens show the measurement settings in single range mode
and dual or auto range mode. In the dual and auto range modes, both the
“High Range” or “Low Range ” buttons averaging buttons will be available.
Home Screen>Settings>Measurement Settings (single range mode)
Home Screen>Settings>Measurement Settings (dual or auto range
mode)
The Measurement Settings screen contains the following information:
Thermo Scientific
●
Averaging Time: Sets the averaging time for the SO2 sample
measurement when in single gas mode.
●
Range Mode Selection: User can choose between the various range
modes: single, dual, or auto. For more information, see “Range Mode
Selection” on page 3-65
●
Range Setting: Sets the concentration range for the analog outputs when
in single range mode.
43iQ Instruction Manual
3-61
Operation
Settings
3-62
43iQ Instruction Manual
●
High Range Setting: Sets the high range concentration range for the
analog output when in dual or auto range mode.
●
Low Range Setting: Sets the low range concentration for the analog
output when in dual or auto range mode.
●
Gas Mode: User can manually choose sample, zero or span mode.
●
Gas Units: Defines the units in which SO2 concentration reading is
expressed.
●
Dilution Ratio: Serves as a multiplier when dilution gas is utilized.
●
Advanced Measurement Settings: Advanced settings affecting SO2
readings.
Thermo Scientific
Operation
Settings
Averaging Time
The Averaging Time screen allows the user to choose dynamic filtering or a
manually selected (static) averaging time.
Averaging Time defines the time period (1 to 300 seconds) during which
SO2 measurements are taken. The average concentration of the readings are
calculated for that time period. The front panel display and analog outputs
are updated every second if averaging time is greater than 1 second. An
averaging time of 10 seconds, for example, means that the average
concentration of the last 10 seconds will be displayed at each update. An
averaging time of 300 seconds means that the moving average
concentration of the last 300 seconds will be the output at each update
(every 10 seconds). Therefore, the lower the averaging time the faster the
front panel display and analog outputs respond to concentration changes.
Longer averaging times are typically used to smooth output data.
Dynamic Filtering allows for data smoothing without compromising
response time. The instrument automatically changes the averaging time
giving the user faster response times when conditions are rapidly changing;
smoother and stable readings, when conditions aren’t changing as rapidly;
and as an added bonus, it better processes spikes to minimize their impact
on the data. At the same time it will preserve the representative nature of
the filtered data to the conditions being sampled.
Note When Dynamic Filtering is selected, the user selected Averaging
Time button is disabled. ▲
Home Screen>Settings>Measurement Settings>Averaging Time (single
range mode)
Thermo Scientific
43iQ Instruction Manual
3-63
Operation
Settings
Home Screen>Settings>Measurement Settings>Averaging Time (dual
or auto range mode)
The Averaging Time screen contains the following information:
3-64
43iQ Instruction Manual
●
Dynamic Filtering: Enables/disables dynamic filtering when in single
range mode.
●
High Range Dynamic Filtering: Enables/disables high range dynamic
filtering when in dual or auto range mode.
●
Low Range Dynamic Filtering: Enables/disables low range dynamic
filtering when in dual or auto range mode.
●
Averaging Time: Sets averaging time period in single range mode and
when dynamic filtering is disabled.
●
High Range Averaging Time: Sets high averaging time when in dual or
auto range mode and when dynamic filtering is disabled.
●
Low Range Averaging Time: Sets low averaging time when in dual or
auto range mode and when dynamic filtering is disabled.
Thermo Scientific
Operation
Settings
Range Mode Selection
The Range Mode Selection screen is used to switch between the various
range modes: Single, Dual, and Auto Range.
Home Screen>Settings>Measurement Settings>Range Mode Selection
The Range Mode Selection screen contains the following information:
●
Single: In single range mode, there is one range, one averaging time,
and one span coefficient.
●
Dual: In the dual range mode, there are two independent analog
outputs. These are labeled simply as the “High Range” and the “Low
Range”. Each channel has its own analog output range setting,
averaging time, and span coefficient.
This enables the sample concentration reading to be sent to the analog
outputs at two different ranges. For example, the low SO2 analog
output can be set to output concentrations from 0 to 50 ppb and the
high SO2 analog output set to output concentrations from 0 to 100
ppb.
In addition, each SO2 analog output has a span coefficient so that each
range can be calibrated separately. This is especially necessary if the two
ranges are not close to one another. For example, the low SO2 range is
set to 0–50 ppb and the high SO2 range is set to 0–10000 ppb.
●
Auto: The auto range mode switches the SO2 analog outputs between
high and low range settings, depending on the concentration level. The
high and low ranges are defined in the Range Settings screen.
For example, suppose the low range is set to 50 ppb and the high range
is set to 100 ppb, as shown below. Sample concentrations below 50
ppb are output based on low range selection and sample concentrations
above 50 ppb are output based on high range selection. When the low
range is active, the range mode selection status output is at 0 volts.
Thermo Scientific
43iQ Instruction Manual
3-65
Operation
Settings
When the high range is active, the range mode selection status output is
at half of full-scale.
When the high range is active, the concentration must drop to 95% of
the low SO2 range for the low range to become active.
In addition, each SO2 range and analog output has a span coefficient so
that each range can be calibrated separately. This is especially necessary
if the two ranges are not close to one another. For example, the low
SO2 range is set to 0–50 ppb and the high SO2 range is set to 0–10000
ppb.
3-66
43iQ Instruction Manual
Thermo Scientific
Operation
Settings
Range Setting
The Range Setting screen defines the concentration range of the analog
outputs. For example, an SO2 range of 0–50 ppb restricts the analog output
to concentrations between 0 and 50 ppb.
The screen shows the current SO2 range. The range screen is similar for the
single, dual, and auto range modes. The only difference between the
screens are the words “High” or “Low” displayed to indicate which range is
displayed. For more information about the dual and auto range modes, see
“Range Mode Selection” on page 3-65. Pressing Range Setting, High
Range Setting or Low Range Setting, brings up a numeric keypad whereby
the user can select a range.
Settable ranges according to unit selection include:
Thermo Scientific
ppb
5010000 ppb
ppm
0.0510 ppm
%
5e-060.001 %
μg/m3
20025000 μg/m3
mg/m3
0.225 mg/m3
g/m3
5e-050.01 g/m3
43iQ Instruction Manual
3-67
Operation
Settings
Gas Mode
The Gas Mode screen defines what gas mode the instrument is set to.
Home Screen>Settings>Measurement Settings>Gas Mode
The Gas Mode screen contains the following information:
3-68
43iQ Instruction Manual
●
Sample: Sets the instrument to measure sample gas.
●
Zero: Used when calibrating the background of the instrument. When
pressed, sets the instrument to zero mode.
●
Span: Used when calibrating the span coefficient. When pressed, sets
the instrument to span mode.
●
External Span: If optional perm oven is installed, allows the use of an
external span source in addition to the internal permeation span.
Thermo Scientific
Operation
Settings
Gas Units
The Gas Units screen defines how the SO2 concentration reading is
expressed. The μg/m3, mg/m3, and g/m3 gas concentration modes are
calculated using a standard pressure of 760 mmHg and a standard
temperature of 0 °C.
Use the and buttons to move up and down.
Home Screen>Settings>Measurement Settings>Gas Units
The Gas Units screen contains the following information:
Thermo Scientific
●
ppb: parts per billion.
●
ppm: parts per million.
●
%: percent.
●
μg/m3: micrograms per meter cubed.
●
mg/m3: milligrams per meter cubed.
●
g/m3: grams per meter cubed.
43iQ Instruction Manual
3-69
Operation
Settings
Advanced Measurement
Settings
The Advanced Measurement Settings screen allows the user to calibrate the
optical bench and set other advanced settings.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings
The Advanced Measurements Settings screen contains the following
information:
3-70
43iQ Instruction Manual
●
Optical Bench Settings: Calibrates detector gain, PMT supply, flash
voltage, and initial flash reference.
●
Flash Lamp: Enables/Disables flash lamp.
●
Optical Span Test: Enables/Disables test LED and reads the LED SO2
concentration.
●
Extended Ranges: Enables/Disables extended ranges feature.
●
Compensation: Allows the user to compensate for changes in
temperature and pressure concentration.
●
Perm Oven Selection: Allows the user to select the operating temperature
of the perm oven.
●
Pressure Calibration: Calibrates pressure.
Thermo Scientific
Operation
Settings
Optical Bench Settings
The Optical Bench Settings screen allows the user to calibrate the optical
bench. The calibration in this screen should be performed if the part
associated with the optical bench has been changed/repaired or if the span
coefficient is outside the limit of 0.5 to 2.0.
Use the and buttons to move up and down and the and
buttons to move left and right.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Optical Bench Settings
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Optical Bench Settings>Continue To Detector Gain Settings
Screen
The Detector Gain Settings screen contains the following information:
●
Thermo Scientific
Across:
●
Description: Define the actions the user can do.
●
Detector Gain: User can manually set or automatically calibrate the
detector gain.
43iQ Instruction Manual
3-71
Operation
Settings
●
●
PMT Supply: User can manually set or automatically calibrate the
photomultiplier tube supply.
●
Lamp Voltage: Displays the current lamp voltage.
●
Lamp Intensity: Displays the current lamp intensity.
Down:
●
Manual Entry: Shows current value. If pressed, user can manually
set the detector gain value.
●
Auto Cal: When pressed, the detector’s auto-calibration process is
initiated. Please allow up to 5 minutes for calibration to complete.
User can stop calibration by pressing the Stop button under
Detector Gain.
●
End Cal: When pressed, the auto calibration is interrupted and the
value does not change.
●
Default Values: When pressed, the default gain for the lamp
intensity detector is restored.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Optical Bench Settings>Continue To PMT Settings Screen
The PMT Settings screen contains the following information:
●
3-72
43iQ Instruction Manual
Across:
●
Description: Defines the actions the user can do.
●
PMT Supply: User can manually adjust the PMT supply value, read
target and calibration concentration, start/stop automatic
calibration, read voltage, frequency and zero offset frequency
values, and reset default values.
●
Current PMT: Displays the current PMT voltage value.
Thermo Scientific
Operation
Settings
●
●
Thermo Scientific
Units: Displays units for each item.
Down:
●
Manual Entry: Shows current value. If the PMT supply cell is
pressed, user can manually change the value.
●
Target Concentration: Displays target concentration.
●
Calibration Concentration: Displays calibration concentration.
●
Auto Calibration: When pressed, the auto-calibration process is
initiated. Please allow up to 5 minutes for calibration to complete.
●
End Calibration: When pressed, the auto calibration is interrupted
and the value does not change.
●
Default Values: When pressed, the default PMT voltage setting is
restored.
43iQ Instruction Manual
3-73
Operation
Settings
Compensation
The Compensation screen provides compensation for any changes to the
instrument’s output signal due to internal instrument temperature, and
pressure variations.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Compensation
The Compensation screen contains the following information:
3-74
43iQ Instruction Manual
●
Temp Compensation: Toggles temperature compensation enabled or
disabled and provides compensation for any changes to the
instrument's output signal due to internal instrument temperature
variations. The effects of internal instrument temperature changes on
the analyzer's subsystems and output have been empirically determined.
This empirical data is used to compensate for any changes in
temperature.
●
Pressure Compensation: Toggles pressure compensation enabled or
disabled and provides compensation for any changes to the
instrument's output signal due to bench pressure variations. The effects
of bench pressure changes on the analyzer's subsystems and output have
been empirically determined. This empirical data is used to compensate
for any change in bench pressure.
Thermo Scientific
Operation
Settings
Pressure Calibration
The Pressure Calibration screen is used to calibrate the pressure sensor to
zero, span, or factory default values.
Use the and buttons to move up and down and the and
buttons to move left and right.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Pressure Calibration
The Pressure Calibration screen contains the following information:
●
●
Across:
●
Description: Lists items in table.
●
Reading: Displays reading of each pressure sensor.
●
Calibration: Starts calibration or resets default values.
Down:
●
●
●
●
Thermo Scientific
Sensor 13 Reading: Under the column labeled Reading, current
reading of each pressure sensor.
Atmospheric Sensor 13: Under the column labeled reading, the user
enters the current atmospheric pressure in mmHg units. Under the
column labelled Calibration, the user presses the Start button to
calibrate the high point of the sensor.
Zero Sensor 13: User should put the pressure sensor under a strong
vacuum. Under the column labeled reading, the user enters the
pressure in mmHg. Under the column labeled Calibration, the user
presses the Start button to calibrate the low point of the sensor.
Reset all values: Resets values to default.
43iQ Instruction Manual
3-75
Operation
Settings
Communications
The Communications screen allows the user to set TCP/DHCP
parameters, Serial settings, Analog I/O and Digital I/O, Email Server, and
Instrument ID. Buttons are grayed out if not selected in
Settings>Configuration.
Home Screen>Settings>More>Communications
The Communications screen contains the following information:
3-76
43iQ Instruction Manual
●
Wired TCP/DHCP: Settings for communicating with the instrument
through wired Ethernet.
●
Serial RS-232/485: Settings for communicating with the instrument
through RS-232/485 protocol. This is only visible if selected in
Settings>Configuration>Communications Board.
●
Analog I/O: Settings for communicating with the instrument through
analog I/O settings. This is only visible if selected in
Settings>Configuration>Analog I/O.
●
Digital I/O: Settings for communicating with the instrument through
digital I/O settings. This is only visible if selected in
Settings>Configuration>Digital I/O.
●
Email Server (SMTP): Settings for communication with email.
●
Instrument ID: Allows the user to edit the instrument identification
number (ID). The ID is used to identify the instrument when using
protocols to control the instrument or collect data. It may be necessary
to edit the ID number if two or more of instruments of the same model
are connected to one computer. Valid instrument ID numbers are from
0 to 127. The 43iQ has a default instrument ID of 1.
Thermo Scientific
Operation
Settings
Wired TCP/DHCP
The Wired TCP/DHCP screen allows the user to communicate with the
instrument via wired TCP/IP settings.
Note When DHCP is enabled, the dynamic IP address is used. When
DHCP is disabled, the static IP address is used. ▲
Home Screen>Settings>More>Communications> Wired TCP/DHCP
(with DHCP enabled)
Home Screen>Settings>More>Communications>Wired TCP/DHCP
(with DHCP disabled)
The Wired TCP/DHCP screen contains the following information:
Thermo Scientific
●
DHCP: Toggles DHCP enabled/disabled.
●
Dynamic IP Address: Dynamic IP address of the instrument.
●
Dynamic Netmask: Dynamic Netmask of instrument.
●
Dynamic Gateway: Dynamic Gateway of instrument.
43iQ Instruction Manual
3-77
Operation
Settings
3-78
43iQ Instruction Manual
●
Static IP Address: Static IP address of the instrument. This is settable
when DHCP is disabled.
●
Static Netmask: Static Netmask of instrument. This is settable when
DHCP is disabled.
●
Static Gateway: Static Gateway of instrument. This is settable when
DHCP is disabled.
●
DNS Server Address: DNS IP address of instrument. This is settable
when DHCP is disabled.
●
Wired MAC Address: Instrument MAC address.
●
Host Name: Host name of instrument.
Thermo Scientific
Operation
Settings
Serial RS-232/485
The Serial RS-232/485 screen allows the user to setup serial
communication. This is only visible if selected in
Settings>Configuration>Communications Board.
Home Screen>Settings>More>Communications>Serial RS-232/485
The Serial RS-232/485 screen contains the following information:
Thermo Scientific
●
Protocol: User selects Modbus or Streaming.
●
Baud Rate: User selectable baud rates from 1200 to 115200.
●
Bits: User selectable between 7 and 8.
●
Parity: User selectable between None, Even, and Odd.
●
Stop Bits: User selectable between 1 and 2.
●
RS 232/485: User selectable between RS-232 and RS-485.
43iQ Instruction Manual
3-79
Operation
Settings
Analog I/O
The Analog I/O screen allows for configuring the analog inputs/outputs.
This is only visible if selected in Settings>Configuration>Analog I/O.
Home Screen>Settings>More>Communications>Analog I/O
The Analog I/O screen contains the following information:
3-80
43iQ Instruction Manual
●
Analog In: Allows the user to view and calibrate voltage inputs from
external devices.
●
Analog Out (Voltage): Allows the user to view voltage outputs from
external devices.
●
Analog Out (Current): Allows the user to view current (mA) outputs.
●
Analog Out Under/Over Range Enabled/Disabled: Allows the user to
select whether or not the analog outputs are allowed to exceed the
selected output range.
Thermo Scientific
Operation
Settings
Digital I/O
The Digital I/O screen allows for configuring the digital inputs/outputs.
This is only visible if selected in Settings>Configuration>Digital I/O.
Home Screen>Settings>Communications>Digital I/O
The Digital I/O screen contains the following information:
Thermo Scientific
●
Digital In: Allows the user to view digital inputs from external devices.
●
Digital Out (Relays): Allows the user to view relay outputs.
●
Digital Out (Solenoids): Allows the user to view solenoid outputs.
●
Advanced Digital I/O: Allows user to test the digital out relays and
solenoids.
43iQ Instruction Manual
3-81
Operation
Settings
Email Server (SMTP)
The Email Server (SMTP) screen allows the user to configure their email
preferences.
Home Screen>Settings>More>Communications>Email Server (SMTP)
The Email Server (SMTP) screen contains the following information:
3-82
43iQ Instruction Manual
●
SMTP Server Address: Address of the user’s email server.
●
From Email Address: The email address that goes in the From field in
emails.
●
SMTP Server Port: Server port of user’s email server.
●
Email Password: Password for SMTP server.
●
Email UserName: User name that is authorized to send email through
SMTP server.
Thermo Scientific
Operation
Settings
Instrument Settings
The Instrument Settings screen allows the user to configure various
instrument settings.
Home Screen>Settings>Instrument Settings
The Instrument Settings screen contains the following information:
Thermo Scientific
●
Display Setup: Sets touch screen display settings.
●
Alarm Setpoints: View and set all available alarm minimum and
maximum values.
●
Language: Read only.
●
Clock: Sets date and time.
●
Pump Power: Manually enables/disables the pump.
43iQ Instruction Manual
3-83
Operation
Settings
Display Setup
The Display Setup allows the user to change the brightness of the display
and choose power save option.
Home Screen>Settings>Instrument Settings>Display Setup
The Display Setup screen contains the following information:
3-84
43iQ Instruction Manual
●
Power Save: Minutes before screen times out. Toggles enable/disable.
●
Power Save Setting: Option whereby the user can display a black screen
after a set amount of inactivity.
●
Brightness: Sets the brightness of the display.
Thermo Scientific
Operation
Settings
Alarm Setpoints
The Alarm Setpoints screen allows the user to view and set all settable
alarm minimum and maximum values.
Note User cannot set alarm limits outside of the acceptable range. The
minimum and maximum alarm limit can also be set by pressing on the
corresponding buttons located in the Settings>Health Check>Status and
Alarms screen. See “Status and Alarms” on page 3-37. ▲
Use the and buttons to move up and down and the and
buttons to move left and right.
Home Screen>Settings>Instrument Settings>Alarm Setpoints
The Alarm Setpoints screen contains the following information:
●
●
Thermo Scientific
Across:
●
Variable: Lists the items that have settable alarm limits.
●
Value: Displays the current value for each item.
●
Low Alarm: User sets low alarm for item.
●
High Alarm: User sets high alarm for item.
●
Units: Units for each item (not settable).
Down:
●
SO2: SO2 concentration alarm.
●
Bkg Check Offset: User can set the maximum allowable background
reading offset for calibration and calibration checks. This is set with
the high alarm only.
43iQ Instruction Manual
3-85
Operation
Settings
3-86
43iQ Instruction Manual
●
Span Check Offset: User can set the maximum allowable span
reading offset for calibration and calibration checks. This is set with
the high alarm only.
●
Chamber Temperature: Chamber temperature alarm.
●
Lamp Intensity: Lamp intensity alarm.
●
Lamp Voltage: Lamp voltage alarm.
●
Instrument Temperature: Instrument temperature alarm.
●
Bench Pressure: Bench pressure alarm.
●
Flow: Flow pressure alarm.
●
Perm Oven Temperature: Perm oven temperature alarm.
Thermo Scientific
Operation
Settings
Clock
The Clock screen allows the user to set the instrument’s date and time,
choose date/time format, time zone, and time server.
Home Screen>Settings>Instrument Settings>Clock
The Clock screen contains the following information:
Thermo Scientific
●
Month
●
Day
●
Year
●
Hours
●
Minutes
●
Seconds
●
Date / Time Parameters: User chooses time zone, timer server, and date
format.
●
Commit: When pressed, date and time are saved.
43iQ Instruction Manual
3-87
Operation
Settings
Date / Time Parameters
The Date / Time Parameters screen allows the user to choose time zone,
time server and date format.
Home Screen>Settings>Instrument Settings>Clock>Date / Time
Parameters
The Date / Time Parameters screen contains the following information:
3-88
43iQ Instruction Manual
●
Time Zone: User selects time zone from table.
●
Time Server Enabled/Disabled: User can enabled/disable the time server
to get periodic clock updates.
●
Date Format: User selects date format.
Thermo Scientific
Operation
Settings
Time Zone
The Time Zone screen allows the user to set the time zone for the Network
Time Protocol (NTP) server. This should be set to the time zone that the
instrument is located in.
Use the and buttons to move up and down.
Home Screen>Settings>Instrument Settings>Clock>Date / Time
Parameters>Time Zone
The Time Zone screen contains the following information:
Thermo Scientific
●
Date Line West(UTC-12)
●
Samoa Time Zone(UTC-11)
●
Aleutian Time Zone(UTC-10)
●
Alaskan Time Zone(UTC-9)
●
Pacific Time Zone(UTC-8)
●
Pacific Daylight Savings(UTC-7)
●
Mountain Time Zone(UTC-7)
●
Mountain Daylight Savings(UTC-6)
●
Central Time Zone(UTC-6)
●
Central Daylight Savings((UTC-5)
●
Eastern Time Zone(UTC-5)
●
Eastern Daylight Savings(UTC-4)
●
Atlantic Time Zone(UTC-4)
●
Mid-Atlantic(UTC-3)
●
South Georgia(UTC-2)
43iQ Instruction Manual
3-89
Operation
Settings
3-90
43iQ Instruction Manual
●
Cape Verde Time(UTC-1)
●
Coordinated Universal Time(UTC-0)
●
Central European Time(UTC+1)
●
Eastern European Time(UTC+2)
●
Further-Eastern European Time(UTC+3)
●
Gulf Standard Time(UTC+4)
●
Yekaterinburg Time(UTC+5
●
Omsk Time(UTC+6)
●
Indochina Time(UTC+7)
●
ASEAN Common Time(UTC+8)
●
Japan Standard Time(UTC+9)
●
Chamorro Time Zone(UTC+10)
●
Sredmnekolymsk Time(UTC+11)
●
New Zealand Standard Time(UTC+12)
Thermo Scientific
Operation
Settings
Time Server
The Time Server screen allows the user to enable/disable the time server to
get periodic clock updates.
Home Screen>Settings>Instrument Settings>Clock>Date / Time
Parameters>Time Server
The Time Server screen contains the following information:
Thermo Scientific
●
Time Server: Enables/Disables periodic clock updates from an NTP
(Network Time Protocol) source.
●
Set Time Server: User can choose specific time server.
●
Set Default: When pressed, default time server will be used.
43iQ Instruction Manual
3-91
Operation
Settings
Date Format
The Date Format screen allows the user to choose from the following
formats: mm/dd/yyyy, dd/mm/yyyy or yyyy-mm-dd.
Use the and buttons to move up and down.
Home Screen>Settings>Instrument Settings>Clock>Date / Time
Parameters>Date Format
The Date Format screen contains the following information:
3-92
43iQ Instruction Manual
●
U.S. Format mm/dd/yyyy
●
European Format dd/mm/yyyy
●
ISO 8601 yyyy-mm-dd
Thermo Scientific
Operation
Settings
Configuration
The Configuration screen allows the user to enable optional features. If an
option is disabled, the corresponding buttons will be grayed out and the
screens will not be available.
Use the and buttons to scroll through the variables. Select the
variables to log by pressing the corresponding cells. Next, press the
Commit Changes button to save selections. Yellow buttons indicate that
the variable is selected. More than one can be chosen.
Home Screen>Settings>Configuration
The Configuration screen contains the following information:
Thermo Scientific
●
Zero/Span Assembly: Enables zero/span valves.
●
Perm Oven: Enables permeation oven option.
●
Analog I/O: Enables analog I/O option.
●
Digital I/O: Enables digital I/O option.
●
Communications Board (RS485): Enables RS485 communication board.
●
Predictive Diagnostics: Enables predictive diagnostics.
43iQ Instruction Manual
3-93
Operation
Settings
Security Access
Levels
The Access Levels screen allows the user to set the instrument to either
View Only or Full Access. When in Full Access, the user will have access to
all screens. When set to View Only, user will not be able to change any
values.
Home Screen>Settings>Security Access Levels (Full Access)
Home Screen>Settings>Security Access Levels (View Only Access)
The Security Access Levels screen contains the following information:
3-94
43iQ Instruction Manual
●
Current Security Access Full Access: Read only. User will be able to
change all values. Password is needed for full access.
●
Current Security Access View Only: Read only. User won’t be able to
change any values. Password is not needed for view only.
●
Change Security Access to View Only: User can switch to view only
mode. Password not needed to change settings to view only access.
●
Change Security Access to Full Access: User can switch to full access
mode. Password is needed to change settings to full access.
Thermo Scientific
Operation
Settings
●
Thermo Scientific
Change Full Access Security Password: Full access password can have a
blank value or user selected password.
43iQ Instruction Manual
3-95
Operation
Settings
Change Security to View
Only Access
The Change Security to View Only Access screen allows the user to set the
instrument to view only.
Home Screen>Settings>Security Access Levels>Change Security Access
to View Only Access
The Change Security to View Only Access screen contains the following
information:
●
Set Access Level to View Only: Programs the instrument to be in the view
only access level, where the user won’t be able to change any values.
●
Cancel: Exit screen.
Note To change security access from view only access to full access, a
keypad will be displayed where the user can enter full access password. ▲
3-96
43iQ Instruction Manual
Thermo Scientific
Operation
Settings
Change Full Access
Security Password
The Change Full Access Security Password screen allows the user to set a
new password for allowing full access.
Home Screen>Settings>Security Access Levels>Change Full Access
Security Password
Home Screen>Settings>Security Access Levels>Change Full Access
Security Password>Continue
Thermo Scientific
43iQ Instruction Manual
3-97
Operation
Settings
The Change Full Access Security Password screens contain the following
information:
3-98
43iQ Instruction Manual
●
Enter Current Security Password: User enters current security password.
●
Continue: Proceeds to next screen.
●
Enter New Security Access Password: User enters new security password.
●
Confirm New Security Access Password: User confirms new security
password for spelling confirmation.
●
Commit New Security Access Password Change: Commits new security
password.
●
Cancel and Return to the Home Screen: Exits screen and returns to the
Home Screen without changing password.
Thermo Scientific
Operation
Settings
USB Drive
The USB Drive screen allows the user to update firmware,
download/upload information, and change the USB password.
Note The USB drive screen only is useable when a USB drive is inserted
into the USB port. When a USB drive is inserted, the user is prompted to
enter the password if a password has been set. ▲
Home Screen>Settings>USB Drive
The USB Drive screen contains the following information:
Thermo Scientific
●
Firmware Update Via USB Drive: If USB is mounted, user can update
instrument firmware.
●
Download Data To USB Drive: User can download/upload
information.
●
Change USB Password: User can change the USB password.
43iQ Instruction Manual
3-99
Operation
Settings
Firmware Update Via
USB Drive
The Firmware Update Via USB Drive screen allows the user to update
instrument firmware from the USB drive.
Home Screen>Settings>USB Drive>Firmware Update Via USB Drive
The Firmware Update Via USB Drive screen contains the following
information:
3-100
43iQ Instruction Manual
●
Update Firmware: User chooses firmware file from USB and updates
instrument firmware. Instrument reboots when update is finished.
●
Exit: User exits without updating firmware.
Thermo Scientific
Operation
Settings
Download Data To USB
Drive
The Download Data To USB Drive screen allows the user to
download/upload data to/from the USB drive.
Home Screen>Settings>USB Drive>Download Data To USB Drive
The Download Data to USB Drive screen contains the following
information:
Thermo Scientific
●
Download Health Check Report: Includes status and alarms, preventive
maintenance, and maintenance history.
●
Download Entire Data Log: Includes the entire data log (from data
logging).
●
Download Service Log: Includes a complete listing of data for all
variables. This is set at the factory.
●
Download System Log: Consists of system log text files, which include a
listing of system errors.
●
Download Calibration History: Includes the data in the calibration
history screen.
●
Download Configuration Data Backup to USB: Allows the user to
download the configuration file from the instrument to the USB.
●
Upload Configuration Data Restore from USB: Allows the user to upload
the configuration files from the USB to the instrument.
●
Restore: Allows the user to upload the configuration files from the USB
to the instrument.
●
Download All Data: Downloads all reports, logs, histories, and backup
information.
43iQ Instruction Manual
3-101
Operation
Settings
Use the following procedure to download data using the USB connection.
1. Plug a flash drive into the USB connection on the front of the
instrument. If a USB password has been previously set, you will be
prompted to enter the USB password to continue. Press Enter to
continue.
2. To continue, select the OK button.
3. The USB Drive will display. Select Download Data To USB Drive.
3-102
43iQ Instruction Manual
Thermo Scientific
Operation
Settings
4. The Download Data to USB Drive screen will display. Select from
various options to download.
5. The instrument will display a “downloading data” message and begin
transferring data to the USB drive.
Note Do not remove the USB drive from the instrument while the data
is downloading. ▲
6. When the data download is complete, the instrument will display a
“Success!” message and display the file name as it is stored on the USB
flash drive. (The file name format is the instrument serial number,
name of download, followed by a date/time stamp.) Remove the USB
flash drive and select the OK button to continue.
Thermo Scientific
43iQ Instruction Manual
3-103
Operation
Settings
Change USB Password
The Change USB Password screen allows the user to set a new password for
accessing USB.
Home Screen>Settings>USB Drive>Change USB Password
Home Screen>Settings>Security Access Levels>Change Standard Access
Password>Continue
3-104
43iQ Instruction Manual
Thermo Scientific
Operation
Settings
The Change USB Password screens contain the following information:
Thermo Scientific
●
Enter Current USB Password: User enters current USB password.
●
Continue: Proceeds to next screen.
●
Enter New USB Password: User enters new USB password.
●
Confirm New Security Access Password: User confirms new security
password for spelling confirmation.
●
Commit New USB Password Change: Commits new USB password.
●
Cancel and Return to the Home Screen: Exits screen and returns to the
Home Screen without changing password.
43iQ Instruction Manual
3-105
Operation
Settings
User Contact
Information
The User Contact Information screen allows the user to enter their contact
information. This is useful when contacting technical support through
emails found at the screen Health Check>File Sharing and Support.
Home Screen>Settings>User Contact Information
The User Contact Information screen contains the following information:
3-106
43iQ Instruction Manual
●
Business Name
●
User Name
●
Alternate User Name
●
User ID
●
Business Address
●
Business Shipping Address
●
To: User Email Address
●
CC: User Email Address 110
●
User Phone Number
●
Alternate User Phone Number
●
Shelter / Lab Phone Number
Thermo Scientific
Operation
Settings
Update Bootloader
The Update Bootloader screen allows the user to update bootloader and
reboot the instrument. If the button is blue, an update to the bootloader is
available. If button is greyed out, no update is needed.
Home Screen>Settings>Update Bootloader
The Update Bootloader screen contains the following information:
Thermo Scientific
●
Reboot and Update: Update bootloader and reboot instrument.
●
Exit: Exits screen.
43iQ Instruction Manual
3-107
Chapter 4
Calibration
This chapter describes the procedures for calibrating the analyzer and
describes the required equipment.
The 43iQ requires initial and periodic calibration according to the
procedures described in this chapter. A quality control plan that allows the
frequency and number of calibration points to be modified depending on
calibration and zero/span check data should be implemented. Such a
quality control program is essential to ascertain the accuracy and reliability
of the air quality data collected. The data compiled for such a program
might include items such as dates of calibration, atmospheric conditions,
control settings and other pertinent data. For more detailed quality
assurance guidelines, see the Quality Assurance Handbook for Air Pollution
Measurement Systems, published by the U.S. EPA, Research Triangle Park,
NC, 27711.
There are a number of conditions which should be met prior to a
calibration or a zero/span check. First, the instrument should have at least
30 minutes to warm up and stabilize. Second, the range used during the
calibration or zero/span check should be the same as that used during
normal monitoring. Third, all operational adjustments to the instrument
should be completed prior to calibration. Fourth, all parts of the gas flow
system, such as sample lines, and particulate filters, which are used in
normal monitoring should also be used during calibration. Finally, it is
recommended that the recording devices and outputs used during normal
monitoring be calibrated prior to the instrument calibration and that they
be used during the calibration or the zero/span check.
Equipment Required
The following equipment is required to calibrate the instrument:
●
Zero air
●
Calibration gas
Note In order to reduce the possibility of interferences during calibration,
all zero gas supplied to the instrument during calibrations should use air as
the zero gas and not nitrogen. The same recommendation also applies to
span gases supplied, where the calibration gas is supplied with a known SO2
concentration with the balance of gas as air and not nitrogen. ▲
Thermo Scientific
43iQ Instruction Manual 4-1
Calibration
Equipment Required
Zero Air Generation
An SO2-free (< 0.0005 ppb) air supply is required for the proper calibration
and checkout of the instrument. There are several methods that are
acceptable to generate this zero air.
Commercial Heatless Air
Dryers
Commercial heatless air dryers filled with a mixed bed of activated charcoal
and a 13X molecular sieve have been found effective in removing SO2 from
compressed air. The use of this type of zero gas system is recommended
when minimum maintenance is of prime importance. This system requires
a source of compressed air. Refer to the manufacturer’s recommendations
for installation of such a system.
Absorbing Column
An absorbing column packed with activated charcoal is acceptable for
scrubbing SO2 from ambient air. Ambient air is forced through a
laboratory gas absorption column packed with the charcoal and the SO2 is
removed to acceptable levels (<0.0005 ppb). The charcoal should be
changed at a minimum of every six months. It may be necessary to change
the charcoal more frequently depending on local conditions.
SO2 Concentration
Standard
A cylinder of SO2 in air containing an appropriate concentration of SO2
suitable for the selected operating range of the analyzer under calibration is
necessary. The assay of the cylinder must be traceable either to a National
Institute of Standards and Technology (NIST) SO2 in Air Standard
Reference Material (SRM) or an NIST/EPA approved gas manufacturer's
Certified Reference Material (CRM).
A recommended protocol for certifying SO2 gas cylinders against a SO2,
SRM or CRM is given in the Quality Assurance Handbook1. The SO2 gas
cylinder should be recertified on a regular basis determined by the local
quality control program.
Calibration Gas
Generation
4-2
43iQ Instruction Manual
A calibration gas system capable of providing accurate levels of SO2
calibration gas between zero and 80% of the full-scale range is required.
The calibration system must provide a flow rate of at least 0.5 LPM for an
instrument with the standard flow (instruments with higher flow rates will
require a higher minimum calibration system flow rate). All calibration gas
should be derived from local or working standards (such as cylinders of
compressed gas or permeation devices) that are certified as traceable to an
NIST primary standard.
Thermo Scientific
Calibration
Equipment Required
Cylinder Gas Dilution
A cylinder gas dilution system, shown in Figure 4–1, can be constructed.
All connections between components in the system should be made with
glass, SS, PTFE or other non-reactive material.
The air flow controller should be capable of maintaining a constant air flow
within ±2% of the required flow rate. The SO2 flow controller should be
capable of maintaining constant SO2 flows within ±2% of the required flow
rate. Ensure both flow controllers are properly calibrated. The pressure
regulator for the standard SO2 cylinder must have a non-reactive
diaphragm and internal parts and a suitable delivery pressure.
External Flow Meter(s)
and Controller(s)
In order to obtain an accurate dilution ratio in the dilution method used
for calibration, the flow rates must be regulated to 1%, and be measured to
an accuracy of at least 2%. The meter and controller can be two separate
devices, or combined in one device. The user's manual for the meter should
be consulted for calibration information.
Additional information on the calibration of flow devices can be found in
the Quality Assurance Handbook1. It should be noted that all flows should
be corrected to 25 °C and 760 mmHg, and that care should be exercised in
correcting for water vapor content.
Figure 4–1. Cylinder Gas Dilution System
Thermo Scientific
43iQ Instruction Manual
4-3
Calibration
Equipment Required
The exact SO2 concentration is calculated from:
[ SO2 ] OUT =
[ SO2 ] STD x F SO2
F D + F SO2
Where:
[SO2]OUT = diluted SO2 concentration at the output manifold, ppm
[SO2]STD = concentration of the undiluted SO2 standard, ppm
FSO2 = flow rate of the SO2 standard corrected to 25 C and 760 mm Hg
FD = flow rate of the dilution air corrected to 25 C and 760 mm Hg
Commercial Precision
Dilution Systems
Commercial precision dilution systems are available which reliably and
accurately dilute a high concentration gas mixture to provide a reliable span
gas. A high concentration (50 ppm) of SO2 in air is precisely diluted to the
concentration range required.
The Thermo Scientific 146iQ Multi-gas Calibrator is one such system for
precision dilution.
Permeation Tube System
Permeation tube systems which precisely maintain a set temperature to
within ±0.1 °C and hold a zero air flow rate to within ±0.5% can be used
for generation of span gas. The flow rate of the permeation system must be
at least 0.5 LPM for proper operation.
A permeation tube system, shown in Figure 4–2 can be constructed. All
connections between components in the system should be made with glass,
PTFE, or other non-reactive material.
The air flow controllers should be capable of maintaining a constant air
flow within ±2% of the required flow rate. Ensure all devices are properly
calibrated and that all flows are corrected to 25 C and 1 atm.
4-4
43iQ Instruction Manual
Thermo Scientific
Calibration
Equipment Required
Figure 4–2. Permeation Tube System
The SO2 output level is calculated from:
[ SO 2 ] OUT =
Where:
PxK
FT
[SO2]OUT = SO2 output concentration in ppm
P = permeation rate in μg/min
FT = total flow rate of gas after mixing chamber (FP + FD) in liters/minute
K (SO2) = 0.382 constant for SO2 permeant
Commercial Permeation
Systems
Commercial permeation systems, such as the Thermo Scientific 146iQ
Multigas Calibration System, are available for this requirement. Refer to
the instrument instruction manual for more information.
In addition to the 146iQ and other commercial permeation systems, the
43iQ can also be configured with an optional permeation oven to supply
calibration gas for the instrument. The operation of the internal
permeation oven is described in Chapter 9, “Optional Equipment”.
Thermo Scientific
43iQ Instruction Manual
4-5
Calibration
Calibration
Calibration
Regulations typically require zero/span calibration when the instrument is
newly installed, moved, repaired, interrupted for more than a few days, or
when span or zero shift by more than 15%.
All gas must be supplied to the instrument at atmospheric pressure. It may
be necessary to employ an atmospheric bypass plumbing arrangement to
accomplish this. Refer to Figure 2–6 in the “Installation and Setup” chapter
for an example of an atmospheric bypass plumbing arrangement.
If a filter is used, all gas must enter the instrument through the filter.
Note The calibration and calibration check duration times should be long
enough to account for the transition (purge) process when switching from
sample to zero and from zero to span. This transition time is the time
required to purge the existing air. ▲
Depending on the plumbing configuration and the instrument, data from
approximately the first minute of a zero calibration or check should be
disregarded because of residual sample air. Also, data from approximately
the first minute of a span calibration or check should be disregarded
because the span is mixing with the residual zero air. ▲
Use the following procedure to perform a zero/span calibration.
1. If the instrument is equipped with the optional zero/span and sample
valves, connect the zero and span gas to the rear panel bulkheads
labeled ZERO IN and SPAN, otherwise connect a source of vented
zero air to the SAMPLE bulkhead. The top right of the title bar
indicates which mode is active: ZERO, SPAN, or SAMPLE.
Note Do not use nitrogen for zero air.
▲
2. To ensure that the zero air is being measured at atmospheric pressure,
check that the zero air flow reported by the analyzer is approximately
0.5 LPM. Actual zero air flow should be 0.8 LPM or more, with the
excess flowing out the atmospheric bypass.
a. From the Home screen choose Settings>Health Check>Status
and Alarms>Flow and Pressure.
b. Make sure that the reported sample flow is less than the actual zero
air flow.
4-6
43iQ Instruction Manual
Thermo Scientific
Calibration
Calibration
3. Monitor the zero air reading and wait for the reading to stabilize.
4. From the Home Screen, press Calibration>Calibrate Background.
The Target Concentration button will read 0.00. The Calculated
Background button will display the background needed to make the
current SO2 concentration go to 0.00.
5. Press Calibrate to set the SO2 reading to zero and to save the new
background.
6. Connect a source of calibration gas to the SAMPLE bulkhead. The
calibration gas concentration should be approximately 80% of the fullscale range.
7. To ensure that the calibration gas is being measured at atmospheric
pressure, check that the sample flow in the analyzer is approximately
0.5 LPM. Actual sample air flow should be 0.8 LPM or more, with the
excess flowing out the atmospheric bypass.
8. Sample the vented calibration gas and wait for the instrument reading
to stabilize.
9. From the Home Screen, choose Calibration>Calibrate Span
Coefficient.
The user sets the span concentration by pressing the Edit Span
Concentration button. The Calculated Span Coefficient button will
show what the span coefficient will be set to if the Calibrate button is
pressed. Pressing the Calibrate button will save the new coefficient and
calibrate the instrument.
10. Enter the SO2 calibration gas concentration using the pushbuttons, and
then press Calibrate to calibrate the instrument to the SO2 calibration
gas.
Note The minimum and maximum analog output configuration values
should be set to the default settings (minimum 0%; maximum 100%). ▲
Thermo Scientific
43iQ Instruction Manual
4-7
Calibration
Calibration in Dual/Auto Range Mode
Calibration in
Dual/Auto Range
Mode
Regulations typically require zero/span calibration when the instrument is
newly installed, moved, repaired, interrupted for more than a few days, or
when span or zero shift by more than 15%.
All gas must be supplied to the instrument at atmospheric pressure. It may
be necessary to employ an atmospheric bypass plumbing arrangement to
accomplish this. Refer to Figure 2–6 in the “Installation and Setup” chapter
for an example of an atmospheric bypass plumbing arrangement.
If a filter is used, all gas must enter the instrument through the filter.
Note The calibration and calibration check duration times should be long
enough to account for the transition (purge) process when switching from
sample to zero and from zero to span. This transition time is the time
required to purge the existing air. ▲
Depending on the plumbing configuration and the instrument, data from
approximately the first minute of a zero calibration or check should be
disregarded because of residual sample air. Also, data from approximately
the first minute of a span calibration or check should be disregarded
because the span is mixing with the residual zero air. ▲
Use the following procedure to perform a zero/span calibration.
1. If the instrument is equipped with the optional zero/span and sample
valves, connect the zero and span gas to the rear panel bulkheads
labeled ZERO IN and SPAN, otherwise connect a source of vented
zero air to the SAMPLE bulkhead. The top right of the title bar
indicates which mode is active: Zero, Span, or Sample.
Note Do not use nitrogen for zero air.
▲
2. To ensure that the zero air is being measured at atmospheric pressure,
check that the zero air flow reported by the analyzer is approximately
0.5 LPM. Actual zero air flow should be 0.8 LPM or more, with the
excess flowing out the atmospheric bypass.
a. From the Home screen choose Settings>Health Check>Status
and Alarms>Flow and Pressure.
b. Make sure that the reported sample flow is less than the actual zero
air flow.
4-8
43iQ Instruction Manual
Thermo Scientific
Calibration
Calibration in Dual/Auto Range Mode
3. Monitor the zero air reading and wait for the reading to stabilize.
4. From the Home Screen, choose Calibration>Calibrate Background.
The Target Concentration button will read 0.000. The Calculated
Background button will display the background needed to make the
current SO2 concentration go to zero.
5. Press Calibrate to set the SO2 reading to zero and to save the new
background.
6. Connect a source of calibration gas to the SAMPLE bulkhead. The
calibration gas concentration should be approximately 80% of the high
full-scale range.
7. To ensure that the calibration gas is being measured at atmospheric
pressure, check that the sample flow in the analyzer is approximately
0.5 LPM. Actual sample air flow should be 0.8 LPM or more, with the
excess flowing out the atmospheric bypass.
8. Sample the vented calibration gas and wait for the instrument reading
to stabilize.
9. From the Home Screen, choose Calibration>Calibrate High Range
Span Coefficient.
The user sets the span concentration by pressing the Edit High Range
Span Concentration button. The Calculated High Span Coefficient
button will show what the high range span coefficient will be set to if
the Calibrate button is pressed. Pressing the Calibrate button will save
the new high range coefficient and calibrate the instrument.
10. Enter the SO2 calibration gas concentration using the pushbuttons, and
then press Calibrate to calibrate the instrument to the high SO2
calibration gas.
Note The minimum and maximum analog output configuration values
should be set to the default settings (minimum 0%; maximum 100%). ▲
11. Connect a source of calibration gas to the SAMPLE bulkhead. The
calibration gas concentration should be approximately 80% of the low
full-scale range.
Thermo Scientific
43iQ Instruction Manual
4-9
Calibration
Zero and Span Check
12. From the Home Screen, choose Calibration>Calibrate Low Range
Span Coefficient.
The user sets the span concentration by pressing the Edit Low range
Span Concentration button. The Calculated Low Range Span
Coefficient button will show what the low span coefficient will be set to
if the Calibrate button is pressed. Pressing the Calibrate button will
save the new low range coefficient and calibrate the instrument.
13. Enter the SO2 calibration gas concentration using the pushbuttons, and
then press Calibrate to calibrate the instrument to the low range SO2
calibration gas.
Zero and Span
Check
The zero and span check procedure is normally performed any time a quick
check of the accuracy of the instrument is required. Normally, zero and
span are checked daily. As experience is gained with the instrument, the
frequency of these checks can be adjusted accordingly.
The span gas concentration used in the span check should be between 70%
and 90% of the full-scale range. The zero and span drift should be
measured and recorded prior to making any adjustments.
All gas must be supplied to the instrument at atmospheric pressure. It will
be necessary to employ an atmospheric bypass plumbing arrangement to
accomplish this.
If a filter is used, all gas must enter the instrument through the filter.
Use the following procedure to check the zero/span.
1. If the instrument is equipped with the optional zero/span and sample
valves, connect the zero and span gas to the rear panel bulkheads
labeled ZERO IN and SPAN, otherwise connect a source of vented
zero air to the SAMPLE bulkhead.
Note Do not use nitrogen for zero air.
▲
2. To ensure that the zero air is being measured at atmospheric pressure,
check that the zero air flow reported by the analyzer is approximately
0.5 LPM. Actual zero air flow should be 0.8 LPM or more, with the
excess flowing out the atmospheric bypass.
a. From the Home screen choose Settings>Health Check>Status
and Alarms>Flow and Pressure.
4-10
43iQ Instruction Manual
Thermo Scientific
Calibration
Manual Calibration
b. Make sure that the reported sample flow is less than the actual zero
air flow.
3. Monitor the zero air reading and wait for the reading to stabilize.
4. Record the measured SO2 value as the zero drift since the last
adjustment. If the zero has changed by more than ±0.015 ppm, it is
recommended that a new calibration be performed.
5. Connect a source of vented span gas to the SAMPLE bulkhead. The
span gas should be approximately 80% of the full-scale range.
6. To ensure that the span gas is being measured at atmospheric pressure
check that the flow is approximately 0.5 LPM or more, with the excess
flowing out the atmospheric bypass.
7. Sample the vented calibration gas and wait for the instrument reading
to stabilize.
8. Record the difference between the measured SO2 value and the actual
SO2 span concentration used. This is the span drift since the last
adjustment. If the calibration has changed by more than ±10%, a new
calibration should be performed.
Manual
Calibration
The Manual Calibration screen allows the user to view and manually adjust
the zero background and span coefficient. These are used to correct the
SO2 readings that the instrument generates using its own internal
calibration data.
Normally, the zero background and span coefficient are calculated
automatically at the Calibrate Background and Calibrate Span Coefficient
described earlier in the chapter. However, the calibration factors can also be
set manually using the functions as described below.
The following screen is shown in single range mode. In dual or auto range
modes, “High Range” or “Low Range” buttons are displayed to indicate the
calibration of the high or low coefficient. The Adjust High Range Span
Coefficient and Adjust Low Range Span Coefficient screens function the
same way.
Thermo Scientific
43iQ Instruction Manual
4-11
Calibration
Manual Calibration
Home Screen>Calibration>Advanced Calibration>Manual Calibration
(single range mode)
Adjust Background
The SO2 background is the amount of signal read by the analyzer while
sampling zero air.
The Adjust Background screen is used to perform a manual zero
background calibration of the instrument. As such, the instrument should
sample zero air until stable readings are obtained. The button labeled
Adjust Background allows the user to change zero background. The second
button called Adjusted Concentration shows what the new SO2
concentration would be based on the changed zero background. Press the
Calibrate button to save the adjusted zero background value.
Home Screen>Calibration>Advanced Calibration>Manual
Calibration>Adjust Background
Adjust Span
Coefficient
The span coefficient is used to correct the SO2 readings and normally has a
value near 1.000 with minimum/maximum limits of 0.500 and 2.000
respectively.
The user can manually change the span coefficient by entering a value in
the Adjust Span Concentration button. The second button called Adjusted
4-12
43iQ Instruction Manual
Thermo Scientific
Calibration
Manual Calibration
Concentration shows what the new SO2 concentration would be based on
the adjusted span coefficient. Press the Calibrate button to save the
adjusted span coefficient value.
Home Screen>Calibration>Advanced>Manual Calibration>Adjust
Span Coefficient (single range mode)
Home Screen>Calibration>Advanced Calibration>Manual
Calibration>Adjust High Range Span Coefficient (dual or auto range
mode)
Reset Bkg to 0.000
and Span Coef to
1.000
Thermo Scientific
The Reset Bkg to 0.000 and Span Coef to 1.000 screen allows the user to
reset the calibration configuration values to factory defaults.
43iQ Instruction Manual
4-13
Calibration
Zero/Span Schedule
Zero/Span
Schedule
The Zero/Span Schedule screen is available only if the zero/span valve
option is installed and turned on (toggles enabled or disabled) at the screen
Settings>Configuration. It is used to program the instrument to perform
fully automated zero and span calibration or calibration checks.
Home Screen>Calibration>Zero/Span Schedule
Home Screen>Calibration>Zero/Span Schedule>More
Next Time
Period
Zero/Span/Purge
Duration Minutes
4-14
43iQ Instruction Manual
The Next Time button is used to view and set the initial date and time (24hour format) of the zero/span schedule. Once the zero/span schedule
begins, the date and time of the next zero/span schedule is calculated and
displayed.
The Period button defines the period or interval between zero/span
schedule. Periods between 0 and 999 hours are acceptable. To turn the
zero/span schedule off, set the period to 0.
The Zero Duration button defines how long zero air is sampled by the
instrument. The Span and Purge Duration buttons look and function the
Thermo Scientific
Calibration
References
same way as the zero duration button. The span duration button is used to
set how long the span gas is sampled by the instrument. The purge
duration button is used to set how long the purge period will be after doing
a zero and/or span. This gives the instrument time to flush out the zero and
span gas before any meaningful sample data is taken. Logged data is flagged
as taken during a purge to show that the data is suspect. Durations between
0 and 99 minutes are acceptable. Each time a zero/span schedule occurs the
zero is done first, followed by the span. To perform just a zero, set the span
duration to 0 (off). The same applies to perform just a span.
Schedule Averaging
Time
Background
Calibration and
Span Calibration
Zero/Span Ratio
References
The Schedule Averaging Time button allows the user to adjust the schedule
averaging time. The schedule averaging time is used by the analyzer only
when performing a zero/span schedule. The analyzer’s averaging time is
used for all other functions. Range is 1300 seconds.
Background Calibration and Span Calibration are toggle buttons that
change between enabled or disabled.
If the background calibration is set to enabled, then a zero adjustment is
made. If the span calibration is set to enabled, then a span adjustment is
made. (This is how to set up a scheduled, recurring auto calibration.)
The Zero/Span Ratio button is used to set the ratio of zero checks or
adjustments to span checks or adjustments. For example, if this value is set
to 1, a span duration will follow every zero duration. If this value is set to 3,
there will be two zero checks between each span check. This value may be
set from 1 to 99, with 1 as default.
1. Section 12 of EPA Quality Assurance Handbook for Air Pollution
Measurement Systems, Volume II, available at www.epa.gov.
Section 12 also provides information on “Calibration of Primary and
Secondary Standards for Flow Measurements”.
Specific information on certification of concentration standards is given
in EPA Traceability Protocol for Assay and Certification of Gaseous
Calibration Standards, available at www.epa.gov.
Thermo Scientific
43iQ Instruction Manual
4-15
Chapter 5
Maintenance
This chapter describes the periodic maintenance procedures that should be
performed on the instrument to ensure proper operation. Since usage and
environmental conditions vary greatly, you should inspect the components
frequently until an appropriate maintenance schedule is determined.
Safety
Precautions
Read the safety precautions before beginning any procedures in this
chapter.
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component. For more
information about appropriate safety precautions, see the “Servicing”
chapter. ▲
Fan Filter Inspection
and Cleaning
Use the following procedure to inspect and clean the fan filter.
1. Remove the fan guard from the fan and remove the filter.
2. Flush the filter with warm water and let dry (a clean, oil-free purge will
help the drying process) or blow the filter clean with compressed air.
3. Re-install the filter and fan guard.
Thermo Scientific
43iQ Instruction Manual 5-1
Maintenance
Pump Rebuilding
Pump Rebuilding
Use the following procedure to rebuild the pump.
Equipment required:
Pump rebuild kit (qty. 1)
Phillips drive, #1 or Torque drive, T10 (depending on pump version)
Pencil or marker
Figure 5–1. Single Stage Pump
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Mark the position of head parts relative to each other by drawing a line
with a pencil. This helps avoid incorrect assembly later.
3. Undo the four screws in the head.
4. Lift the head plate and the intermediate plate off the housing.
5. Hold the pump with one hand, so that the diaphragm is pointing
downwards. Lift the diaphragm by the opposing side edges, grasp it and
unscrew it in the counter-clockwise direction.
5-2
43iQ Instruction Manual
Thermo Scientific
Maintenance
Pump Rebuilding
6. Remove connection rod disc and diaphragm spacers from the threaded
pin of the diaphragm.
7. Push the connection rod disc and the diaphragm spacers in this order
onto the threaded pin of the new diaphragm.
8. Move the connecting rod to the upper point.
9. Screw the new diaphragm with connection rod disc and spacers
clockwise onto the connection rod and tighten hand-tight.
10. Place the intermediate plate on housing, in the position indicated by
the drawing line.
11. Place the new valve plate on the intermediate plate.
12. Place the head plate on the intermediate plate, in the position indicated
by the drawing line; gently tighten the four screws, evenly and
diagonally (if a torque screwdriver is available: torque about 0.30 Nm).
13. Let the pump run.
Thermo Scientific
43iQ Instruction Manual
5-3
Maintenance
Pump Rebuilding
Figure 5–2. Pump Rebuilding
5-4
43iQ Instruction Manual
Thermo Scientific
Maintenance
Leak Test
Leak Test
Use the following procedure to perform a leak test.
Equipment Required:
Cap
Vacuum Tester with Gauge (with a resolution of .5 in Hg or better)
1. Turn instrument OFF, unplug the power cord.
2. Block the SAMPLE bulkhead on the rear panel with a leak-tight cap.
3. Connect the vacuum tester tool to the EXHAUST bulkhead on the
rear panel.
4. Squeeze trigger until gauge reads to pull in 10 in Hg.
5. Observe vacuum gauge for stable reading for 5 minutes. If reading
remains at 10 in Hg, no leak is present.
Note Acceptable leak rate is .5 in Hg over 10 minutes. ▲
Thermo Scientific
43iQ Instruction Manual
5-5
Maintenance
Lamp Voltage Check
Lamp Voltage
Check
The instrument is equipped with a lamp voltage control circuit, which
automatically corrects for degradation of the flash lamp. However, after
several years of use, the lamp may have degraded to the point that is being
driven with the maximum voltage that the power supply can deliver.
There are two methods available that allows the operator to check the lamp
voltage. The first is via the Measurements Settings>Advanced Measurement
Settings>Optical Bench Settings>Continue to Detector Gain Settings
Screen. The current Lamp Voltage and Lamp Intensity are displayed.
The second method is via Settings>Health Check>Status and Alarms>SO2
Bench screen. The current Lamp Voltage and Lamp Intensity are displayed
along with the upper and lower alarm limits.
For detailed information about this screen, refer to the “Operation”
chapter. For more information about replacing the lamp or adjusting the
lamp voltage control circuit, see the “Servicing” chapter.
Lamp Voltage
Adjustment
The auto calibration of the detector will automatically adjust the lamp
voltage to 800 volts (±5 Volts). This is done so by adjusting the gain on the
detector to optimize the illumination of the lamp.
Use the following procedure to adjust the lamp voltage.
1. From the Home screen choose Settings>Measurement
Settings>Advanced Measurement Settings>Optical Bench
Settings>Continue To Detector Gain Settings Screen.
2. Under the Detector Gain – Auto Cal, Press “Start” to initiate the auto
calibration sequence.
5-6
43iQ Instruction Manual
Thermo Scientific
Maintenance
Lamp Voltage Adjustment
3. To manually adjust the detector gain, touch the numeral value under
the detector gain. Then change the Manual Entry Detector Gain to
your set value and save.
Thermo Scientific
43iQ Instruction Manual
5-7
Chapter 6
Troubleshooting
This chapter presents guidelines for diagnosing analyzer failures, isolating
faults, and includes recommended actions for restoring proper operation.
Safety Precautions
Troubleshooting
Guide
Read the safety precautions in Appendix A, “Safety” before performing any
actions listed in this chapter.
Table 6–1 provides general troubleshooting information for the common
platform and indicates the checks that you should perform if you
experience an instrument problem. It also lists 43iQ specific
troubleshooting information and alarm messages you may see on the
graphics display and provides recommendations about how to resolve the
alarm condition.
Table 6–1. 43iQ Troubleshooting Guide
Problem
Instrument does not start
(LEDs on front panel do not
come on and display is
blank)
Thermo Scientific
Possible Cause
Action
No power
Verify that the power cord is
plugged in, power is available and
that it matches the voltage and
frequency configuration of the
instrument.
Fuse is blown or missing
Disconnect power and check
fuses with a volt meter.
Bad switch or wiring
connection to switch
Check for 24V @ J9 on the
Backplane board (middle pins).
Check all wiring connections.
Front panel display does
not start (LEDs on front
panel are off)
Disconnected ribbon
cable
Power down and evaluate
connections of display ribbon
cable.
Front panel display does
not start (LEDs on front
panel are on)
Defective Display
Connect to the instrument using
ePort. Select “Remote Interface”.
If normal GUI is displayed, replace
defective display.
Front panel display stays
white after power up (LEDs
on front panel are on)
Unseated or missing
Micro SD card
Power off, re-seat Micro SD or
install if missing.
43iQ Instruction Manual 6-1
Troubleshooting
Troubleshooting Guide
Problem
Possible Cause
Action
Micro SD Card
Programming
If Micro SD card was just
replaced, re-install the old one. If
the problem is fixed, request a
replacement Micro SD card.
Solenoid current out of
range (option)
Sticking or damaged
solenoid
Reset solenoid via
Settings>Health Check>Status
and Alarms>Valve and Pump
Resets screen. If damaged,
replace solenoid valve block.
Pump current out of range
Damaged or dirty pump
Reset pump via Settings>Health
Check>Status and Alarms>Valve
and Pump Resets. Inspect and
refurbish pump. If pump motor is
damaged, replace pump.
Cannot zero instrument or
there is a high background
signal when sampling zero
air. (Zero air should produce
a reading equivalent to less
than 15 ppb SO2.)
Zero air system is faulty,
needs new SO2 scrubbers
or requires maintenance.
Test against an ultra-zero cylinder
from a reputable scientific gas
supplier or check effect of a new
chromatography grade activated
charcoal scrubber installed at the
instrument inlet.
Zero air flow rate is
inadequate
Check by-pass or atmospheric
pressure vent to verify that the
zero air system is providing more
flow than the instrument is
drawing.
Instrument is not
drawing in zero air
Check the sample flow and
pressure in the Settings>Health
Check>Status and Alarms>Flow
and Pressure screen.
Use an independent flow meter to
check flows at the sample inlet
and exhaust bulkheads (they
should match).
Perform a leak test as described
on page 5-5.
6-2
43iQ Instruction Manual
Span gas containing SO2,
NO or hydrocarbon is
contaminating system
Verify that span gases connected
to the calibration system are shut
off and leak-tight.
Internal or external lines,
filters and other sample
handling equipment are
contaminated or dirty
Replace inlet filter (if installed)
and as much as possible.
Hydrocarbon kicker has
failed
Replace the hydrocarbon kicker.
High scattered light
Toggle the “Flash Lamp” off in the
Settings>Measurement
Thermo Scientific
Troubleshooting
Troubleshooting Guide
Problem
Instrument appears to zero,
but there is weak or no
response to span gas
Possible Cause
Action
Settings>Advanced
Settings>Flash Lamp screen. If the
previously high signal drops to
zero or less when the flash lamp
is off, the problem may be caused
by scattered light from dust in the
optical bench. If so, carefully
clean the bench.
Span cylinder empty
Check the source pressure.
Calibration system
failure
Check solenoids or other
hardware to be sure that span gas
is being delivered.
Flow rate of the diluted
span mix is adequate
Check by-pass or atmospheric
vent to verify that the zero air
system is providing more flow
than the instrument draws.
Instrument is not
drawing in span gas
Check the sample flow and
pressure in the Settings>Health
Check>Status and Alarms>Flow
and Pressure screen.
Use an independent flow meter to
check flows at the sample inlet
and exhaust bulkheads (they
should match).
Perform a leak test as described
on page 5-5.
Zero or Span will not
stabilize
Thermo Scientific
SO2 is being absorbed by
tubing, filters or dirt in
the calibration system
Replace any lines made of vinyl or
plastics with fresh PTFE or
stainless steel. Replace PTFE filter
membranes that look dirty.
Remove any filters that are not
PTFE membranes.
Flash lamp has failed
Listen for the rapid clicking of the
flash lamp. Check the flash lamp
voltage and intensity.
PMT has failed
Check the PMT voltage and run
the optical span test. A good
optical span test indicates that
the PMT is OK and that the
problem is more likely the lamp.
Adjust PMT voltage if needed.
Flow rate of the diluted
span mix is inadequate
Check the by-pass or atmospheric
pressure vent to verify that the
zero air system is providing more
flow than the instrument is
43iQ Instruction Manual
6-3
Troubleshooting
Troubleshooting Guide
Problem
Possible Cause
Action
drawing.
Instrument is not
drawing in span gas
Check the sample flow and
pressure in the Settings>Health
Check>Status and Alarms>Flow
and Pressure screen.
Use an independent flow meter to
check flows at the sample inlet
and exhaust bulkheads (they
should match).
Perform a leak test as described
on page 5-5.
Reduced response or no
response to sample gas
with alarm(s) indicated
SO2 is being absorbed
and released by dirt in
tubing or filters of the
calibration system or
contamination inside the
instrument
Replace any lines made of vinyl or
plastics with fresh PTFE or
stainless steel. Replace PTFE filter
membranes that look dirty.
Remove any filters that are not
PTFE membranes.
Averaging time is not
correct
Check the “Averaging Time” in
the Settings>Measurement
Settings screen. If too high, the
unit will be slow to stabilize. If
too low, the signal may appear to
be noisy.
Flasher lamp
Replace with known-good lamp to
see if the lamp is the problem.
Undefined electronic
failure or pump failure
Check the status and alarms
screen to localize fault.
Check the response to known
span gas.
Run an optical span test.
Instrument is not
drawing in sample as
expected
Check the sample flow and
pressure in the Settings>Health
Check>Status and Alarms>Flow
and Pressure screen.
Use an independent flow meter to
check flows at the sample inlet
and exhaust bulkheads (they
should match).
Perform a leak test as described
on page 5-5.
Reduced response or no
response to sample gas
with no alarm(s) indicated
6-4
43iQ Instruction Manual
Instrument is not
properly calibrated
Verify the SO2 background and
SO2 coefficient are set
appropriately in the Calibration
screen.
Thermo Scientific
Troubleshooting
Troubleshooting Guide
Problem
Possible Cause
Instrument is not
drawing in sample as
expected
Action
Check sample flow and pressure
readings in the Settings>Health
Check>Status and Alarms>Flow
and Pressure screen.
Use an independent flow meter to
check flows at the sample inlet
and exhaust bulkheads (they
should match).
Perform a leak test as described
on page 5-5.
Check the external plumbing for
leaks or other problems.
Check all external plumbing and
the source of the sample to verify
that the SO2 is not being absorbed
by the sampling system. Lines
carrying SO2 must be made from
clean PTFE or stainless steel.
Span calibration coefficient
outside acceptable limits of
0.5–2.0.
Excessive noise or spikes
on analog outputs
Detection circuit failure
Run the optical span text to verify
PMT and associated electronics.
PMT failure
Check the PMT voltage. Adjust if
needed.
Flash lamp failure
Check the lamp voltage. Perform a
detector Autocal. If outside of
spec, replace the detector.
Bad span gas
Verify quality of span gas.
System leak
Perform a leak test as described
on page 5-5.
Insufficient calibrator
flow
Verify calibrator is providing a
flow of at least 0.8 LPM.
Defective or low
sensitivity
Check PMT voltage and run an
optical span test. Replace PMT
with known-good unit if possible.
Noise pick-up by recorder
or data logger
Check analog cable shielding and
replace.
Try to localize source of noise by
comparing analog signal to data
collected through RS-232 or
Ethernet.
Poor linearity
Thermo Scientific
Problem with calibrator
Verify accuracy of the calibration
system with an independent flow
meter.
43iQ Instruction Manual
6-5
Troubleshooting
Troubleshooting Guide
Problem
Excessive response time
Possible Cause
Action
Hydrocarbon kicker
Replace kicker if older than 3
years.
Leak in sample probe line
Check for variable dilution.
Averaging time is not set
correctly
Check the “Averaging Time” in
the Settings>Measurement
Settings screen.
Instrument is not
drawing in sample at the
expected flow rate
Check sample flow and pressure
readings in the Settings>Health
Check>Status and Alarms>Flow
and Pressure screen.
Use an independent flow meter to
check flows at the sample inlet
and exhaust bulkheads (they
should match).
Perform a leak test as described
on page 5-5.
SO2 is being absorbed
and released by dirt in
tubing or filters of the
calibration system or
contamination inside the
instrument
Replace any lines made of vinyl or
plastics with fresh PTFE or
stainless steel.
Replace PTFE filter membranes
that look dirty.
Remove any filters that are not
PTFE membranes.
Analog signal doesn’t
match expected value
Flow meter fluctuations
6-6
43iQ Instruction Manual
Contaminated bench
Contact technical support for
additional information and
options.
Software has not been
configured
Verify that the selected analog
output has been properly
configured to match the data
system.
Analog output goes
above full-scale value or
below zero
By default, a 5% over and under
range on the analog outputs is
provided. If this is not desirable
due to system restrictions, it may
be turned off in the
Settings>Communications>Analog
I/O screen.
Recorder is drawing
down output
Verify that the recorder or data
logger input impedance meets
minimum requirements.
Dirty pump diaphragm
Clean or replace pump diaphragm.
Thermo Scientific
Troubleshooting
Troubleshooting Guide
Problem
Alarm – Chamber
Temperature
Alarm – Perm Gas Temp
Alarm – Pressure
Possible Cause
Action
Capillary blocked
Clean or replace capillary.
Clogged PTFE line
Inspect all sample lines.
Heater failed
Replace heaters as needed.
Defective PCB
Replace PCB.
Incorrect Perm oven set
temperature or alarm
setting
Check that the alarm settings
match the set temperature.
Perm oven heater
thermistor or gas
thermistor error
Replace thermistor.
Perm oven failure
Replace the Perm oven.
High pressure indication
Check plumbing for leaks.
Check the pump for a tear in the
diaphragm. Replace if needed.
Check that the capillary is
properly installed and o-rings are
in good shape.
Alarm – Flow
Flow low
Check sample capillary for
blockage. Replace as necessary.
If using sample particulate filter,
make sure it is not blocked.
Disconnect sample particulate
filter from the sample bulkhead. If
flow increases, replace the filter.
Perform a leak test as decribed on
page 5-5.
Alarm – Intensity
Flow high
When delivering zero air or gas to
the instrument, use an
atmospheric dump.
Flow = 0 LPM
Check that Step POL board #1 has
both dip switch settings of SW2
off (both facing the rear of the
instrument). Verify the pump is
plugged into the Step POL board.
Worn Diaphragm
Rebuild pump every 12 months or
as needed.
Low – Lamp is failing
Check that the lamp and trigger
pack are securely fastened.
Replace lamp.
Alarm – Lamp voltage
Thermo Scientific
Low voltage (<750v)
Replace lamp power supply.
High voltage (>1200v)
Replace flash lamp.
43iQ Instruction Manual
6-7
Troubleshooting
Troubleshooting Guide
Problem
Alarm – SO2 Conc.
Check to ensure range
corresponds with expected value.
If not, select proper range.
Concentration low
Check user-defined low set point.
Be sure the minimum trigger is set
as desired.
Cable connection
Check that DMC cable is
connected properly. Reseat if
needed.
Defective DMC PCB
Replace DMC board.
Cable connection
Check that DMC cable is
connected properly. Reseat if
needed.
Defective component
Check for other alarms, as it is
possible that another component
of that DMC is drawing too much
current.
Defective DMC PCB
Replace DMC board.
Cable connection
Check that DMC cable is
connected properly. Reseat if
needed.
Other alarm
Make sure the instrument
temperature is not too high or in
alarm.
Defective DMC PCB
Replace DMC board.
Alarm – 5V/24V Step Board
Cable connection
Check the cable connections to
that Step POL board.
Alarm – Internal
Temperature
Fam failure
Replace fan if not operating
properly..
Dirty fan filter
Clean or replace filter.
Overheating PCB
Locate defective PCB reporting
the error and replace if needed.
Alarm – Analog I/O
Defective PCB
Replace Analog board.
Alarm – Digital I/O
Defective PCB
Replace Digital board.
Alarm – Auto Bkg
Cal/Check
Incorrect high alarm limit
Verify the high limit is correct via
Settings>Status and
Alarms>Concentrations screen.
Instrument background
calibration failed
Recalibrate the instrument.
Incorrect high alarm limit
Verify the high limit is correct via
Settings>Status and
Alarms>Concentrations screen.
Alarm – Power Supply
Alarm – Module
Temperature
Alarm – Auto Span
Cal/Check
43iQ Instruction Manual
Action
Concentration has
exceeded range limit
Alarm – Board
Communication
6-8
Possible Cause
Thermo Scientific
Troubleshooting
Troubleshooting Guide
Problem
Possible Cause
Instrument span
calibration failed
Thermo Scientific
Action
Recalibrate the instrument.
43iQ Instruction Manual
6-9
Chapter 7
Servicing
This chapter describes the periodic servicing procedures that should be
performed on the instrument to ensure proper operation and explains how
to replace the 43iQ subassemblies.
Safety
Precautions
Read the safety precautions before beginning any procedures in this
chapter.
Icon Here
The service procedures in this manual are restricted to qualified service
representatives. ▲
Icon Here
If the equipment is operated in a manner not specified by the
manufacturer, the protection provided by the equipment may be
impaired. ▲
CAUTION If the LCD panel breaks, do not let the liquid crystal contact
your skin or clothes. If the liquid crystal contacts your skin or clothes, wash
immediately using soap and water. ▲
Do not remove the LCD panel or frame from the LCD module. ▲
The LCD polarizing plate is very fragile, handle it carefully. ▲
Do not wipe the LCD polarizing plate with a dry cloth, as it may easily
scratch the plate. ▲
Do not use alcohol, acetone, MEK or other Ketone based or aromatic
solvents to clean the LCD module, but rather use a soft cloth moistened
with a naphtha cleaning solvent. ▲
Do not place the LCD module near organic solvents or corrosive gases. ▲
Do not shake or jolt the LCD module. ▲
Thermo Scientific
43iQ Instruction Manual 7-1
Servicing
Safety Precautions
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component. For more
information about appropriate safety precautions, see “Safety”. ▲
Note If an antistatic wrist strap is not available, be sure to touch the
instrument chassis before touching any internal components. When the
instrument is unplugged, the chassis is not at earth ground. ▲
Figure 7–1. Properly Grounded Antistatic Wrist Strap
Note Ground to unpainted case or outlet as shown. ▲
7-2
43iQ Instruction Manual
Thermo Scientific
Servicing
Firmware Updates
Firmware
Updates
New versions of the instrument software are periodically made available
over Ethernet, USB flash drive, or company website at:
www.thermofisher.com
For more information on installing new firmware, see “Installing New
Firmware” in the iQ Series Communications manual.
Replacement
Parts List
For a complete list of spare parts, visit the company website at:
www.thermofisher.com/43iQ
Refer to Figure 7–2 and Figure 7–3 to identify the component location.
Thermo Scientific
43iQ Instruction Manual
7-3
Servicing
Replacement Parts List
Figure 7–2. 43iQ Component Layout Top View
Figure 7–3. 43iQ Component Layout Side View
7-4
43iQ Instruction Manual
Thermo Scientific
Servicing
Fuse Replacement
Fuse
Replacement
Use the following procedure to replace the fuses.
1. Turn instrument OFF and unplug the power cord.
2. Remove fuse drawer, located on the AC power connector.
3. If either fuse is blown, replace both fuses.
4. Insert fuse drawer and reconnect power cord.
Figure 7–4. Replacing the Fuses
Thermo Scientific
43iQ Instruction Manual
7-5
Servicing
Filter Replacement
Filter
Replacement
Use the following procedure to replace the filter.
1. Turn instrument OFF and unplug the power cord.
2. Starting with top right corner, pull out to remove fan cover.
Figure 7–5. Start with Top Right Corner of Fan Cover
3. Replace filter and snap fan cover back in place.
Figure 7–6. Removing the Fan Cover
7-6
43iQ Instruction Manual
Thermo Scientific
Servicing
Fan Replacement
Fan Replacement
Use the following procedure to replace the fan.
Equipment required:
Phillips drive, #2
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure 2–1).
2. Unplug the fan cable J18.
Figure 7–7. Unplugging the Fan Cable
3. Starting with top right corner, pull out to remove fan cover.
4. Unhook the four latches of the fan cover.
5. Unfasten the four 6-32 screws from the fan housing.
6. Replace fan and reassemble in reverse order.
Thermo Scientific
43iQ Instruction Manual
7-7
Servicing
Fan Replacement
Figure 7–8. Replacing the Fan
7-8
43iQ Instruction Manual
Thermo Scientific
Servicing
Measurement Side Removal and Replacing
Measurement
Side Removal
and Replacing
Use the following procedure to remove and replace the measurement side if
necessary.
Equipment required:
Phillips drive, #2
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure 2–1).
2. Unplug the fan cable J18 (Figure 7–9).
Figure 7–9. Unplugging the Fan Cable
3. Unplug DMC cable (Figure 7–10).
Thermo Scientific
43iQ Instruction Manual
7-9
Servicing
Measurement Side Removal and Replacing
Figure 7–10. Unplugging the DMC Cable
4. Gripping from the top corners of the front panel and pull outwards.
5. Remove three 8-32 flat head screws (Figure 7–11).
Figure 7–11. Unfasten Hardware Front for Measurement Side Removal
7-10
43iQ Instruction Manual
Thermo Scientific
Servicing
Measurement Side Removal and Replacing
6. Swing arm open.
7. Unfasten captive hardware.
8. Remove two 8-32 flat head screws.
9. Pull measurement side out.
10. Replace in reverse order.
Figure 7–12. Unfasten Hardware Rear for Measurement Side Removal
Thermo Scientific
43iQ Instruction Manual
7-11
Servicing
LCD Module Replacement
LCD Module
Replacement
Use the following procedure to replace the LCD module.
Equipment required:
Wrench, 1/4
1. Turn instrument OFF and unplug the power cord.
2. Gripping from the top corners of the front panel and pull outwards.
3. Unfasten four nuts (Figure 7–13).
Figure 7–13. Replacing the LCD Module
4. Remove cover.
5. Unplug LCD cables from backside of board.
6. Pull board off the standoffs.
7-12
43iQ Instruction Manual
Thermo Scientific
Servicing
LCD Module Replacement
Figure 7–14. Remove Electrical Cables from LCD
7. Replace LCD module and reassemble in reverse order.
Thermo Scientific
43iQ Instruction Manual
7-13
Servicing
I/O Replacement
I/O Replacement
Use the following procedure to replace the I/O boards.
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure 2–1).
2. Swing arm open.
Figure 7–15. I/O Replacement, Arm
3. Pull board upwards.
Figure 7–16. I/O Replacement, Remove Board
7-14
43iQ Instruction Manual
Thermo Scientific
Servicing
I/O Replacement
4. During install, make sure to align cutout circular to keyway.
5. Insert board downwards.
Figure 7–17. I/O Replacement, Install
6. Close arm. Make sure expansion bracket aligns to the inside of the
rectangular cutouts.
Figure 7–18. I/O Replacement, Close Arm Alignment
Thermo Scientific
43iQ Instruction Manual
7-15
Servicing
Peripherals Support Board and System Controller Board Replacement
Peripherals
Support Board
and System
Controller Board
Replacement
Use the following procedure to replace the peripherals support board or
system controller board.
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure 2–1).
2. Pull tab out (two per board).
3. Pull board out.
Figure 7–19. Replacing the Peripherals Support Board
4. Replace board and reassemble in reverse order.
7-16
43iQ Instruction Manual
Thermo Scientific
Servicing
DMC Pressure and Flow Board
DMC Pressure
and Flow Board
Use the following to replace the DMC pressure and flow board.
Equipment required:
Hex drive, 7/16
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure 2–1).
2. Unplug cables from the pressure and flow board.
Figure 7–20. Flow Pressure Board, Disconnect DMC Cable
3. Disconnect plumbing.
4. Using 7/16 hex drive, unfasten four #6-32 socket cap head screws.
Thermo Scientific
43iQ Instruction Manual
7-17
Servicing
DMC Pressure and Flow Board
Figure 7–21. Flow Pressure Board, Screws
5. Replace board and reassemble in reverse order.
7-18
43iQ Instruction Manual
Thermo Scientific
Servicing
Pump Replacement
Pump
Replacement
Use the following procedure to replace the pump.
Equipment required:
Phillips drive, #1 and #2
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure 2–1).
2. Unplug pump cable from step POL board J7.
3. Twist opposite direction to unlock tube clamps.
Note Push in tube clamp to lock. ▲
4. Disconnect tubing from pump.
5. Unfasten two captive hardware.
6. Slide pump left until keyway meets opening.
Thermo Scientific
43iQ Instruction Manual
7-19
Servicing
Pump Replacement
Figure 7–22. Remove Pump, Disconnect and Unfasten
7. Pull pump outwards.
Note When installing pump, make sure the pump keyway opening goes
over the keyway. ▲
7-20
43iQ Instruction Manual
Thermo Scientific
Servicing
Pump Replacement
Figure 7–23. Pump Removal, Keyway
8. Remove two screws.
Figure 7–24. Pump replacing, Unfasten Screws
9. Replace pump and reassemble in reverse order.
Thermo Scientific
43iQ Instruction Manual
7-21
Servicing
Capillary Cleaning and/or Replacement
Capillary
Cleaning and/or
Replacement
Use the following procedure to clean or replace the capillary.
Equipment required:
Phillips drive, #2
Hex drive, 7/64
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure 2–1).
2. Disconnect the plumbing.
3. Unfasten captive hardware.
Figure 7–25. Remove Capillary, Disconnect and Unfasten
4. Slide capillary plate upwards clearing the partition panel keyway.
7-22
43iQ Instruction Manual
Thermo Scientific
Servicing
Capillary Cleaning and/or Replacement
Figure 7–26. Capillary Plate, Keyway
5. Using 7/64 hex drive, remove four #6-32 socket cap head screws.
6. Pull apart the capillary blocks.
Figure 7–27. Capillary and O-Ring Replace
Thermo Scientific
43iQ Instruction Manual
7-23
Servicing
Capillary O-Ring Replacement
7. Replace capillary and reassemble in reverse order.
Note Fitting arrangements, number of capillaries and capillary sizes will
vary per instrument configuration. ▲
Capillary O-Ring
Replacement
Use the following procedure to replace the capillary o-rings.
Equipment required:
O-ring pick tool
1. Using a plastic o-ring pick tool, remove the o-ring.
Note Be careful in not damaging the o-ring walls during this process. Refer
to Figure 7–27. ▲
Power Supply
Replacement
Use the following procedure to replace the power supply.
Equipment required:
Phillips drive, #2
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure 2–1).
2. Unplug all electrical shown J9, J10, J24, J25, J26, and ground.
3. Unfasten captive hardware.
4. Slide power supply left, clearing three case floor plate keyways.
7-24
43iQ Instruction Manual
Thermo Scientific
Servicing
Power Supply Replacement
Figure 7–28. Removing Power Supply
5. Pull power supply up.
6. Replace power supply and reassemble in reverse order.
Figure 7–29. Replacing Power Supply
Thermo Scientific
43iQ Instruction Manual
7-25
Servicing
Step POL Board Replacement
Step POL Board
Replacement
Use the following procedure to replace the Step POL Board.
Equipment required:
Torque screwdriver, T15 or Slot screwdriver, 3/16
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure 2–1).
2. Unplug step POL power cable J4.
3. Unplug step POL signal cable J2.
4. Unplug pump cable J7.
5. Unfasten captive hardware.
Figure 7–30. Unplug and Unfasten Step POL Board
7-26
43iQ Instruction Manual
Thermo Scientific
Servicing
Step POL Board Replacement
Figure 7–31. Clear Partition Keyway Step POL Board
6. Slide step board 1 upwards clearing the partition panel keyway.
7. If replacing step board 1, make sure switch 1 and 2 are pointed away
from ON (Figure 7–32). If replacing optional step board 2, make sure
switch 1 is pointed towards ON and switch 2 is pointed away from ON
(Figure 7–33).
8. Replace step POL board and reassemble in reverse order.
Thermo Scientific
43iQ Instruction Manual
7-27
Servicing
Step POL Board Replacement
Figure 7–32. Step POL Board 1 Switch Settings
Figure 7–33. Optional Step POL Board 2 Switch Settings
7-28
43iQ Instruction Manual
Thermo Scientific
Servicing
DMC Optical Bench Removal
DMC Optical
Bench Removal
Optical Bench
Assembly Removal
Use the following to remove the DMC optical bench from the instrument.
●
Optical Bench Assembly Removal
●
PMT Replacement
●
Flasher Pack Replacement
●
Flasher Pack Lamp Replacement
●
Photo Lamp Detector Board Replacement
●
Kicker Assembly Replacement
Use the following procedure to remove the optical bench assembly.
Equipment required:
Phillips drive, #2
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure 2–1).
2. Unplug two DMC cables.
3. Disconnect all plumbing.
4. Unfasten six captive hardware.
Thermo Scientific
43iQ Instruction Manual
7-29
Servicing
DMC Optical Bench Removal
Figure 7–34. Removing the DMC Optical Bench Removal from Instrument
7-30
43iQ Instruction Manual
Thermo Scientific
Servicing
DMC Optical Bench Removal
Optical Bench
Assembly Removal
Use the following to remove the optical bench assembly.
Equipment required:
Hex nut drive, 3/8 and 5/32
1. Unplug all electrical as shown (Figure 7–35).
Figure 7–35. Removing the Optical Bench Assembly
2. Unfasten 10-32 nut.
3. Unfasten four 10-32 screws.
4. Pull optical bench assembly upwards (Figure 7–36).
Thermo Scientific
43iQ Instruction Manual
7-31
Servicing
DMC Optical Bench Removal
Figure 7–36. Removing the Optical Bench Assembly pt 2
7-32
43iQ Instruction Manual
Thermo Scientific
Servicing
DMC Optical Bench Removal
Photomultiplier
Tube (PMT)
Replacement
Use the following to replace the photomultiplier tube (PMT).
Equipment required:
Hex drive, 5/32
Phillips drive, #1 and #2
Lint-free gloves
1. Turn the instrument OFF, unplug the power cord, and remove the
cover.
2. Unplug all electrical as shown (Figure 7–37).
3. Unfasten two 10-32 screws.
4. Pull photomultiplier housing upwards.
Figure 7–37. Removing the PMT Assembly
Thermo Scientific
43iQ Instruction Manual
7-33
Servicing
DMC Optical Bench Removal
5. Unfasten three 4-40 screws.
Figure 7–38. Removing the PMT Cover
6. Slide cover and grommet back.
Note Wear lint-free gloves to unplug PMT from base. ▲
7. Unfasten two 6-32 screws.
8. Replace PMT as needed and assemble in reverse order.
Figure 7–39. Replacing the PMT
7-34
43iQ Instruction Manual
Thermo Scientific
Servicing
DMC Optical Bench Removal
Flasher Pack
Replacement
Use the following procedure to replace the flasher pack.
Equipment required:
Hex drive, 1/16
1. Unplug flash trig from board.
2. Loosen 6-32 set screw until the flasher pack/lamp can be eased out of
housing.
Figure 7–40. Removing and Replacing the Flasher Pack
Thermo Scientific
43iQ Instruction Manual
7-35
Servicing
DMC Optical Bench Removal
Flasher Pack Lamp
Replacement
Use the following procedure to replace and flasher pack lamp.
Equipment required:
Lint-free gloves
1. Wearing lint-free gloves unplug the lamp from the flasher pack.
Figure 7–41. Replacing the Flasher Pack Lamp
2. Replace flasher pack lamp and assemble in reverse order.
7-36
43iQ Instruction Manual
Thermo Scientific
Servicing
DMC Optical Bench Removal
Photo Lamp
Detector Board
Replacement
Use the following procedure to remove and replace the photo lamp
detector board.
Equipment required:
Phillips drive, #1 and #2
1. Unplug cable.
2. Unfasten three 6-32 screws.
Figure 7–42. Replacing the Photo Lamp Detector Board pt 1
3. Pull detector cover outwards.
4. Unfasten four 4-40 screws.
5. Pull photo lamp detector board outwards.
6. Replace lamp detector board and assemble in reverse order.
Thermo Scientific
43iQ Instruction Manual
7-37
Servicing
DMC Optical Bench Removal
Figure 7–43. Replacing the Photo Lamp Detector board pt 2
7-38
43iQ Instruction Manual
Thermo Scientific
Servicing
DMC Optical Bench Removal
Kicker Assembly
Replacement
Use the following procedure to replace the kicker assembly (Figure 7–44).
Equipment required:
Phillips drive, #1
1. Disconnect plumbing as shown.
2. Loosen 4-40 screw. Slide kicker assembly towards opening of keyway.
Equipment Damage Do not loosen or tighten this side of fitting as the
internal tubing will twist blocking flow. These fittings are covered by
silicone tape as a reminder. ▲
3. Once kicker assembly clears the keyway, pull outwards towards the
board and up.
Figure 7–44. Removing and Replacing the Kicker Assembly
Thermo Scientific
43iQ Instruction Manual
7-39
Servicing
Optional Manifold Replacement
4. Replace the kicker assembly and assemble in reverse order.
Optional
Manifold
Replacement
Use the following procedure to replace the manifold.
Equipment required:
Hex wrench, 9/16
Hex drive, 9/64
1. Turn the instrument OFF, unplug the power cord, and remove the
cover (Figure 2–1).
2. Unplug three electrical connections (J5, J6, and J8) from the step POL
board 1.
Figure 7–45. Replacing the Manifold pt 1
3. Unfasten three nuts. Remove the nuts, front and back ferrules as shown
from span, zero in, sample back panel (Figure 7–46).
7-40
43iQ Instruction Manual
Thermo Scientific
Servicing
Optional Manifold Replacement
Figure 7–46. Replacing the Manifold pt 2
4. Unfasten fitting nut and slide tee fitting assembly off.
5. Unfasten four #8-32 screws.
Figure 7–47. Replacing the Manifold pt 3
6. Replace the manifold and assemble in reverse order.
Thermo Scientific
43iQ Instruction Manual
7-41
Servicing
Optional DMC Permeation Oven Solenoid Valve Replacement
Optional DMC
Permeation Oven
Solenoid Valve
Replacement
Use the following procedure to replace the DMC permeation oven solenoid
valve option.
Equipment required:
Phillips drive, #1
CAUTION Allow the oven to cool down prior to servicing. ▲
1. Turn the instrument OFF, unplug the power cord, and remove the
cover (Figure 2–1).
2. Unplug perm valve J7 from step pol board 2.
Figure 7–48. Removing and Replacing the Solenoid Valve pt 1
3. Loosen two 4-40 screws and slide optional perm valve out.
7-42
43iQ Instruction Manual
Thermo Scientific
Servicing
Optional DMC Permeation Oven Solenoid Valve Replacement
Equipment Damage Do not disconnect the plumbing from the valve end.
Disconnect from the attaching end only. This will prevent damaging and
leaks from the valve end. ▲
Figure 7–49. Removing and Replacing the Solenoid Valve pt 2
4. Replace the permeation oven solenoid valve and assemble in reverse
order.
Thermo Scientific
43iQ Instruction Manual
7-43
Servicing
Permeation Oven Replacement
Permeation Oven
Replacement
Use the following procedure to replace the permeation oven.
Equipment required:
Phillips drive, #2
Tube release tool (optional)
CAUTION Allow oven to cool down prior to servicing. ▲
1. Turn the instrument OFF, unplug the power cord, and remove the
cover (Figure 2–1).
2. Unplug DMC cables as shown.
3. Disconnect plumbing.
Figure 7–50. Perm Oven pt 1
7-44
43iQ Instruction Manual
Thermo Scientific
Servicing
Permeation Oven Replacement
4. Push in fitting head evenly with fingers towards fitting body.
Figure 7–51. Finger Push and Release Tubing
5. Pull tube outwards to release. If using the tool provided:
a. Adjust tool size to 1/4 tube as indicated on the back side.
b. Tool edge should be pushed straightforward into the edge of the
release button along with the tube in the axial fitting direction.
c. After inserting, grasp handle tightly and insert the end of the tubing
to the stroke end.
Note Insert firmly to the guard against an accidental tube release. ▲
d. After inserting end of tube, relax your grip on the tool. Returning
force of spring releases the tube.
Thermo Scientific
43iQ Instruction Manual
7-45
Servicing
Permeation Oven Replacement
Push both sides at once to release. Reverse and fix at the same position as
before. Applicable tube size is indicated on the back side.
6. Disconnect plumbing from inline fitting.
Note Do not disconnect from valve end. ▲
7. Unfasten two captive hardware.
Figure 7–52. Perm Oven Replacement pt 2
8. Replace DMC permeation oven and assembly in reverse order.
7-46
43iQ Instruction Manual
Thermo Scientific
Servicing
Permeation Oven Board Replacement
Permeation Oven
Board
Replacement
Use the following procedure to replace the permeation oven board.
Equipment required:
Phillips drive, #2
CAUTION Allow oven to cool down prior to servicing. ▲
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure 2–1).
2. Unplug all electrical J1, J7, J8, J9 as shown.
3. Using #2 Phillips drive, unfasten four 6-32 pan head screws.
Figure 7–53. Replacing the Permeation Oven Board
4. Replace permeation board and assemble in reverse order.
Thermo Scientific
43iQ Instruction Manual
7-47
Chapter 8
System Description
The 43iQ deploys a set of modular subsystems that comprise the total
instrument function. The core measurements for concentration are
contained in Distributed Measurement and Control (DMC) modules. This
chapter describes the function and location of the system components in
the module framework, including firmware, electronics, and I/O function.
The 43iQ system components include:
●
●
Heated hydrocarbon kicker
●
Optical Bench hardware with bandpass filters and mirrors
●
Flash lamp trigger assembly
●
Photomultiplier tube (PMT)
●
Photodetector
●
Optical bench DMC board
●
Permeation Oven (optional)
●
Common Electronics
●
●
Thermo Scientific
Optical Bench DMC
●
Power Supply
●
System controller board
●
Backplane board
●
Front panel
●
I/O (optional)
Peripherals Support System
●
Fan (on rear panel)
●
STEP POL board
●
Sample pump
●
Solenoid valve panel (optional)
Flow Pressure DMC with flow restricting capillary
43iQ Instruction Manual 8-1
System Description
Optical Bench DMC
●
Optical Bench
DMC
The Optical Bench DMC contains the key components of the optical
measurement that eventually results in SO2 concentration. In the reaction
chamber pulsating light from the flash lamp excites the SO2 molecules. A
condenser lens collects and focuses light from fluorescing SO2 molecules
onto the PMT assembly.
Heated Hydrocarbon
Kicker
The heated hydrocarbon kicker removes hydrocarbons from the gas stream
while leaving the SO2 concentration unaffected. It operates on a selective
permeation principle using differential pressure to force hydrocarbon
molecules to pass through the tube wall. The differential pressure is created
across the tube wall as sample gas passes through a capillary tube which
reduces its pressure. The sample gas is then fed to the shell side of the
hydrocarbon kicker. The heated kicker operates at the same temperature as
the optical bench and requires no additional power.
Optical Bench
Hardware
The optics section provides the light source for the fluorescence reaction
and optimizes the reaction with a system of lenses and mirrors. It includes a
flash lamp, condensing lens, bandpass mirror assembly and light baffle.
Condensing Lens
The condensing lens focuses light from the flash lamp into the mirror
assembly.
Mirror Assembly
A set of four mirrors selectively reflects only those wavelengths used in
exciting SO2 molecules. This reflective filtering causes the radiation
reaching the detection chamber to be more intense and more stable
throughout the lifetime of the instrument.
Light Baffle
Bandpass Filter
8-2
Firmware
43iQ Instruction Manual
The circular baffle helps keep stray light from entering the detection
volume.
The bandpass filter restricts the light reaching the photomultiplier tube to
the SO2 fluorescence wavelengths.
Thermo Scientific
System Description
Optical Bench DMC
Flash Lamp Trigger
Assembly
The flash lamp trigger assembly pulses the UV flash lamp at a rate of 10
times per second for improved signal-to-noise ratio and long term stability.
The lamp is operated in the pulsed mode for six major reasons:
Flash Lamp
Flash Trigger Board
●
Long life
●
High optical intensity – improved signal to noise ratio
●
Small size
●
Low power requirements – less than 1 watt
●
Long-term stability
●
Chopped signal processing – no dark current drift
The flash lamp provides the ultraviolet light source that causes the
fluorescence reaction in the SO2 molecules.
The flash trigger board is located in the base of the flash lamp assembly. It
receives high voltage and the trigger signal from the measurement interface
board and uses a small transformer to produce a short, high-voltage pulse to
fire the flash lamp.
Flash Intensity Assembly
The flash intensity assembly is located in the fluorescence chamber and
continuously monitors the pulsating UV flash lamp. This photodetector is
connected to a circuit that automatically compensates for fluctuations in
flash lamp output.
Photomultiplier
Tube
The PMT power supply produces high voltage to operate the
photomultiplier tube used in the measurement system. The output voltage
is adjustable from approximately 600 to 1200 volts under software control.
The PMT converts optical energy from the reaction to an electrical signal.
This signal is sent to the input board which transmits it to the processor.
Optical Bench DMC
Board
The optical bench DMC board accepts the current signal from the PMT
and converts it to a voltage, which is scaled by a factor of approximately 1,
10, or 100 depending on the full-scale range of the SO2 channel.
Thermo Scientific
43iQ Instruction Manual
8-3
System Description
Permeation Oven (Optional)
Bench Heater
The fluorescence chamber temperature is measured with a thermistor. The
voltage across the thermistor is fed to the main processor for use in
calculating and displaying the reaction chamber temperature. The voltage
across the thermistor is also compared to a set-point voltage and used to
control that the reaction chamber heaters to maintain a constant
temperature.
Permeation Oven
(Optional)
The 43iQ can be configured with an optional permeation oven for
generating SO2 span gas. The permeation oven is configured as a DMC
module with self-contained temperature measurements and heater controls.
For more information, see “Internal Permeation Span Source” as described
in Chapter 9, “Optional Equipment”.
Common
Electronics
The common electronics contain the core computational and power
routing hardware for the 43iQ, and is replicated throughout other iQ series
products (Figure 8–1). It also contains the front panel display, the USB
ports, the Ethernet port, and the optional I/O interfaces (RS-485, analog,
and digital).
Figure 8–2 shows the PCBA interconnect structure for the 43iQ, including
options. The modular design of the instrument is conveyed in the
architecture. Brief descriptions of the specific PCBAs follow.
8-4
43iQ Instruction Manual
Thermo Scientific
Thermo Scientific
MFC 1
MFC 2
MFC 3
ANALOG I/O
BOARD
GND
POL 1, 24V
POWER
POL 1
DATA
BOARD
DMC 1
COMMUNICATIONS
Not all connections apply to all Model iQ instruments.
+24V Fan
Measurement Case
Backplane
LAN
USB
DIGITAL I/O
BOARD
LINE POWER FILTER
PERIPHERAL'S
24V
POWER 1
DMC 2
SUPPORT BOARD
POL 2
DATA
POL 2, 24V
POWER
BACKPLANE
BOARD
RESET
24V
POWER 2
A/C
OUT
FRONT USB PORT BOARD
DISPLAY
BOARD
48VDC
24VDC
A/C
OUT
FRONT PANEL
SWITCH
A/C IN
GROUND TAB
SYSTEM
CONTROLLER BOARD
"B"
"A"
-
iQ Series Common Platform
FRONT PANEL
LEDS
TOUCH SCREEN LCD DISPLAY
+48V POWER SUPPLY
(Only if 48V is required)
+24V POWER SUPPLY
System Description
Common Electronics
Figure 8–1. Common System Interconnect Diagram
43iQ Instruction Manual
8-5
8-6
43iQ Instruction Manual
Sample
Sol
Pump
Fan
Zero Sol
Span Sol
Measurement Case
Backplane
24V
Fan
POL 1, 24V
POWER
Trigger Pack
!
POL 1
DATA
HVPS
FLASH TRIG
Step Pol
Module 1
43 PMT DMC
PMT
Step Pol
Module 2
Perm Oven Sol
Flow Pressure
DMC
DMC 1
HEATER
LED
LED
Heater
Heater
iQ Series 43
Permeation Oven DMC
Sensor
DMC 2
System Description
Common Electronics
Figure 8–2. 43iQ System Interconnect Diagram
Thermo Scientific
System Description
Common Electronics
Power Supply
Front Panel
I/O and
Communication
Components
All electronics operate from a universal VDC supply, which is capable of
auto-sensing the input voltage and working over all specified operating
ranges. The 43iQ contains a 24 VDC channel for most electronics
operation, including the pump and fan, and a 48 VDC channel dedicated
specifically for optical bench heating.
Front panel electronic components include the touch screen display, the on
off switch, and two indicator LEDs for power and alarm status, as
described in operational detail in Chapter 2, “Installation and Setup”.
The iQ series instruments provide a number of methods for
communicating the instrument results to the operator or external
equipment. Every iQ series instrument includes a front panel display, 2
USB ports, and one Ethernet data port as standard equipment.
In addition, optional RS-232/485, analog I/O, and digital I/O ports are
available to provide data to external systems as described in Chapter 9,
“Optional Equipment”. The front panel GUI allows the operator to
configure these output communication channels as described in Chapter 3,
“Operation”.
System Controller
Board
The System Controller Board (SCB) contains the main processor, power
supplies, and a sub-processor, and serves as the communication hub for the
instrument. The SCB receives operator inputs from the front panel GUI
and/or over optional I/O connections on the rear panel. The SCB sends
commands to the other boards to control the functions of the instrument
and to collect measurement and diagnostic information. The SCB outputs
instrument status and measurement data to the GUI, Ethernet/USB, and
to the optional rear-panel I/O. The SCB plugs into the backplane via a
single connector and has physical retainers to secure placement.
Backplane Board
The backplane board provides the routing and conditioning for +24 VDC
(optional +48 VDC) and RS-485 communications within the instrument.
It hosts the System Controller Board (SCB) and Peripherals Support Board
(PSB) via direct plug ins, and similarly hosts optional I/O (communication,
analog, and digital) with rear panel interfaces via direct plug in. It has
connections for RS485 communication with and powering of DMCs and
the STEP POL Module. It additionally routes the front panel display and
driver, external USB and Ethernet.
Thermo Scientific
43iQ Instruction Manual
8-7
System Description
Peripherals Support System
Peripherals
Support System
The peripheral support system operates these additional devices that are
needed, but do not require special feedback control or processing. These
components are connected to a Peripherals Support Board (PSB).
Fan
The chassis fan provides air cooling of the active electronic components.
STEP POL Board
Sample Pump
Solenoid Valve
Panel
Flow/Pressure
DMC
The STEP POL board provides high/low outputs for continuous operation
or on/off states. The STEP POL board contains the basic circuitry to
provide a programmable load to passive devices, either continuously, or on
user or automated command. In the iQ Series instruments, the pump,
solenoids, etc., are controlled off of the STEP POL board from commands
generated via the PSB.
Internal vacuum pump for generating air/sample through the instrument
Optional solenoid valves for switching between sample, zero, and span
gases, and other optional components.
The Flow/Pressure DMC is used measure instrument pressures that assure
proper flow regulation and for sample pressure within the measurement
bench for pressure corrections and compensation.
The DMC includes two pressure sensors that read 0-860 mmHg. These
sensors are used with the coupled restricting capillary for flow control along
with the downstream sample pump. The pressure differential determines
the flow through the capillary. The upstream pressure is the measurement
bench pressure, while the downstream pressure is at the pump inlet
pressure.
Firmware
8-8
43iQ Instruction Manual
Like the hardware, the firmware is modular and located within
microprocessors distributed throughout the instrument. In the 43iQ,
microprocessors containing firmware are located as follows:
●
Optical Bench DMC
●
Flow/Pressure DMC
●
Peripherals Support Board
●
Optional I/O (Communications, Digital, and Analog)
Thermo Scientific
System Description
Firmware
●
Optional Permeation Oven
The firmware contains the active controls for their application, as well as
self-identification and configuration for “plug and play” style operation.
Each are associated with specific registers of two types:
●
Modbus registers that are communicated from each microprocessor to
the System Controller Board (SCB) via internal RS-485
●
SNMP registers that are maintained in the software and SCB for health
and data processing computation
The Modbus communication system operates on 1 second intervals.
Within those intervals, data treatment like integration (whether analog or
digital) and servo control, are embedded in the module firmware. The SCB
receives the 1 second updates for higher level “software” processing and
control via SNMP registers, some of which is interfaced with the front
panel Graphical User Interface (GUI).
In addition to the operating registers, the 43iQ stores a historical data log
in a MySQL database. The memory is provided on the same μSD card
where the operating software resides, for which there is capability to store
up to a year of data at 1 minute intervals. Chapter 3, “Operation” describes
how this database is accessed and used including external memory
downloads.
Thermo Scientific
43iQ Instruction Manual
8-9
Chapter 9
Optional Equipment
The 43iQ is available with the following options:
Connecting
External Devices
Several components are available for connecting external devices.
These connection options consist of three plug-in boards:
●
Communication Board
●
Analog I/O Board
●
Digital I/O Board
Figure 9–1. I/O Expansion Replacement Boards
Thermo Scientific
43iQ Instruction Manual 9-1
Optional Equipment
Connecting External Devices
Communication
Board
RS-232/RS-485 Port
The communication board consists of:
●
RS-232/485 Port
●
RS-485 External Accessory Port
The RS-232/RS-485 port uses a 9-pin serial connector with a bi-directional
serial interface that can be configured for either RS-232 or RS-485
communication.
Figure 9–2. RS-232/RS-485 Port
Table 9–1. RS-232/RS-485 Port Terminal Assignment
9-2
43iQ Instruction Manual
Terminal Number
Signal Name
1
No Connect
2
RX/RS485_RX_P
3
TX/RS485_TX_N
4
No Connect
5
GND
6
No Connect
7
RTS/RS485_TX_P
8
CTS/RS485_RX_N
9
No Connect
Thermo Scientific
Optional Equipment
Connecting External Devices
RS-485 External
Accessory Port
The RS-485 external accessory port uses a 15-pin serial connector for
communication with external smart devices.
Figure 9–3. RS-485 External Accessory Port
Table 9–2. RS-485 External Accessory Port Terminal Assignment
Thermo Scientific
Terminal Number
Signal Name
1
EXT_RS485_RX_N
2
EXT_RS485_RX_P
3
+5V (Fused @0.4A)
4
+5V (Fused @0.4A)
5
+5V (Fused @0.4A)
6
GND
7
GND
8
GND
9
EXT_RS485_TX_N
10
EXT_RS485_TX_P
11
+24V (Fused @0.4A)
12
+24V (Fused @0.4A)
13
+24V (Fused @0.4A)
14
+24V (Fused @0.4A)
15
+24V (Fused @0.4A)
43iQ Instruction Manual
9-3
Optional Equipment
Connecting External Devices
Analog I/O Board
Analog Voltage Inputs
The Analog I/O Board consists of:
●
4 Isolated Analog Voltage Inputs, Input Voltage Range: 0–10 V
●
6 Isolated Analog Voltage Outputs, Three Ranges: 0–1.0 V, 0–5.0 V,
0–10 V
●
6 Isolated Analog Current Outputs, Two Ranges: 0mA–20mA, 4mA–
20mA
Table 9–3 lists the analog voltage inputs are used to monitor four external
0–10V signals.
Figure 9–4. Analog Voltage Inputs
Table 9–3. Analog Voltage Inputs Assignment
9-4
43iQ Instruction Manual
Terminal Number
Signal Name
1
Analog In 1
2
Analog GND
3
Analog In 2
4
Analog GND
5
Analog In 3
6
Analog GND
7
Analog In 4
8
Analog GND
Thermo Scientific
Optional Equipment
Connecting External Devices
Analog Voltage Outputs
There are six globally isolated, 16-bit, Analog Output channels, each with a
Voltage Output, a Current Output and a common Return (isolated
ground). The Analog Outputs are configured through the software control
registers to select Voltage Output ranges 0–1 V, 0–5 V or 0–10 V, as well
as Current Output ranges 0–20 mA or 4–20 mA. The maximum allowable
load for each Current Output is 1000 Ω. All Voltage Outputs and Current
Outputs are continuously monitored separately for accuracy.
The Analog Outputs may be used to control and report parameters
pertinent to the analyzers’ measured functions.
Figure 9–5. Analog Voltage and Current
Thermo Scientific
43iQ Instruction Manual
9-5
Optional Equipment
Connecting External Devices
Table 9–4. Analog Voltage and Current Assignment
Analog Output
Calibration
Terminal Number
Signal Name
1
Current Out 1
2
Voltage Out 1
3
C/V Return 1
4
Current Out 2
5
Voltage Out 2
6
C/V Return 2
7
Current Out 3
8
Voltage Out 3
9
C/V Return 3
10
Current Out 4
11
Voltage Out 4
12
C/V Return 4
13
Current Out 5
14
Voltage Out 5
15
C/V Return 5
16
Current Out 6
17
Voltage Out 6
18
C/V Return 6
The iQ series instruments provide for the ability to calibrate the analog
outputs (both voltage and current) of the instruments. The basic procedure
for both voltage and current are the same using the following procedure:
●
Complete the connections of the recording device to the desired analog
output channel. (See page 9-5 for the channel information).
●
Calibrate the output channel low level.
Note When calibrating the current output when using the 0-20 mA
scale, the low level will be set to 4 mA due to the inability to adjust the
actual current output to below zero. ▲
●
9-6
43iQ Instruction Manual
Calibrate the output channel full scale.
Thermo Scientific
Optional Equipment
Connecting External Devices
Analog Output Zero
Calibration
Use the following procedure to calibrate the output channel to low level.
This analog output calibration procedure reflects the zero calibration for
analog output voltage for demonstration purposes. To calibrate the 4 mA
current calibration, follow the same procedure, by selecting the 4 mA
current calibration option.
Note This adjustment should only be performed by an instrument service
technician. ▲
1. From the Home screen, choose Settings>Communications>Analog
I/O>Analog Out Calibration.
2. Depending on the output type being used, select either Analog Out
Zero Calibration (Voltage) or Analog Out 4.000 mA Calibration
(Current).
3. A confirmation screen is presented. Select Continue to proceed with
the calibration or Return to Previous Screen.
Thermo Scientific
43iQ Instruction Manual
9-7
Optional Equipment
Connecting External Devices
4. There are six columns for each of the six available output channels:
●
●
●
●
Output (V): Displays the actual output level at the terminal of the
analog output board. For analog voltage, this value will default at zero.
For analog current, this value will default at 4 mA.
Decrease and Decrease : Decreases the output by coarse or fine
amounts
Increase and Increase : Increases the output by coarse and fine
amounts.
Commit: Accepts the changes to the analog output levels.
5. For the desired analog output channel, increase or decrease the output
until the reading on the recording device indicates the proper value.
6. After making changes to the output levels, the commit button will turn
green. To accept the changes, press the Commit button. To revert to
the previous values, press the back button to return to the previous
analog output calibration screen.
9-8
43iQ Instruction Manual
Thermo Scientific
Optional Equipment
Connecting External Devices
Analog Output Full Scale
Calibration
Use the following procedure to calibrate the output channel to full scale.
This analog output calibration procedure reflects the full scale calibration
for analog output voltage for demonstration purposes. To calibrate the 20
mA current calibration, follow the same procedure, by selecting the 20 mA
current calibration option.
Note This adjustment should only be performed by an instrument service
technician. ▲
1. From the Home screen, choose Settings>Communications>Analog
I/O>Analog Output Calibration.
2. Depending on the output type being used, select either Analog Out
Full Scale Calibration (Voltage) or Analog Out 20.000 mA Calibration
(Current).
3. A confirmation screen is presented. Select Continue to proceed with
the calibration or Return to Previous Screen.
Thermo Scientific
43iQ Instruction Manual
9-9
Optional Equipment
Connecting External Devices
4. There are six columns for each of the six available output channels:
●
●
●
●
Output (V): Displays the actual output level at the terminal of the
analog output board. For analog voltage, this value will default at the
setting of the output channel, 1, 5, or 10 V. For analog current, this
value will default at 20 mA.
Decrease and Decrease : Decreases the output by coarse or fine
amounts
Increase and Increase : Increases the output by coarse and fine
amounts.
Commit: Accepts the changes to the analog output levels.
5. For the desired analog output channel, increase or decrease the output
until the reading on the recording device indicates the proper value.
6. After making changes to the output levels, the commit button will turn
green. To accept the changes, press the Commit button. To revert to
the previous values, press the back button to return to the previous
analog output calibration screen.
9-10
43iQ Instruction Manual
Thermo Scientific
Optional Equipment
Connecting External Devices
Digital I/O Board
Digital Inputs
The digital I/O board consists of:
●
16 Digital Inputs (18 pin connector)
●
10 Digital Relay Switches (20 pin connector)
●
8 Valve Driver Outputs (16 pin connector)
The digital inputs are TTL (3 V or 5 V) compatible and are pulled high
within the instrument. The active state can be user defined in firmware.
●
Logic Low Threshold: 0.8 V
●
Logic High Threshold: 2.0 V
●
Absolute allowable input voltages: -0.5 to 5.5 V
Figure 9–6. Digital Inputs
Thermo Scientific
43iQ Instruction Manual
9-11
Optional Equipment
Connecting External Devices
Table 9–5. Digital Inputs Terminal Assignment
Terminal Number
Signal Name
COMMON
1
Digital In 1
2
Digital In 2
3
Digital In 3
4
Digital In 4
5
Digital In 5
6
Digital In 6
7
Digital In 7
8
Digital In 8
9
Digital In 9
10
Digital In 10
11
Digital In 11
12
Digital In 12
13
Digital In 13
14
Digital In 14
15
Digital In 15
16
Digital In 16
COMMON
9-12
43iQ Instruction Manual
Thermo Scientific
Optional Equipment
Connecting External Devices
Digital Relay Switches
Table 9–6 lists the digital relay switches.
●
Maximum Voltage: 300 VDC
●
Maximum Current: 500 mA
●
Fuse: 800 mA
Figure 9–7. Digital Relay Switches
Thermo Scientific
43iQ Instruction Manual
9-13
Optional Equipment
Connecting External Devices
Table 9–6. Digital Relay Switch Assignment
9-14
43iQ Instruction Manual
Terminal Number
Signal Name
1A
Relay 1A
1B
Relay 1B
2A
Relay 2A
2B
Relay 2B
3A
Relay 3A
3B
Relay 3B
4A
Relay 4A
4B
Relay 4B
5A
Relay 5A
5B
Relay 5B
6A
Relay 6A
6B
Relay 6B
7A
Relay 7A
7B
Relay 7B
8A
Relay 8A
8B
Relay 8B
9A
Relay 9A
9B
Relay 9B
10A
Relay 10A
10B
Relay 10B
Thermo Scientific
Optional Equipment
Connecting External Devices
Valve Driver Outputs
Table 9–7 lists the valve driver outputs.
●
Actual Output Voltage: 2224 VDC
●
Maximum Current: 300 mA
●
Both positive and negative outputs are protected from over voltage and
over current by 500 mA fuses.
Figure 9–8. Valve Driver Outputs
Thermo Scientific
43iQ Instruction Manual
9-15
Optional Equipment
Connecting External Devices
Table 9–7. Valve Driver Outputs Assignment
Terminal Number
Signal Name
1+
Valve Drive 1+
1-
Valve Drive 1-
2+
Valve Drive 2+
2-
Valve Drive 2-
3+
Valve Drive 3+
3-
Valve Drive 3-
4+
Valve Drive 4+
4-
Valve Drive 4-
5+
Valve Drive 5+
5-
Valve Drive 5-
6+
Valve Drive 6+
6-
Valve Drive 6-
7+
Valve Drive 7+
7-
Valve Drive 7-
8+
Valve Drive 8+
8-
Valve Drive 8-
Note Intended for 24 V valves. These outputs will also drive any DC load
of 2224 VDC, up to 300 mA. ▲
9-16
43iQ Instruction Manual
Thermo Scientific
Optional Equipment
Internal Zero/Span and Sample Valves
Internal Zero/Span
and Sample Valves
With the zero/span assembly option, a source of span gas is connected to
the SPAN port and a source of zero air is connected to the ZERO IN port.
It may be necessary to use an atmospheric dump bypass plumbing
arrangement to accomplish this.
For more information, refer to the “Installation and Setup” chapter and the
“Operation” chapter.
Internal
Permeation Span
Source
The Internal Permeation Span Source option is designed to provide a
simple span check. This option is intended as a quick, convenient check to
be used between zero and span calibrations for determining instrument
malfunction or drift. Because this option does not precisely control dilution
gas flow, it should not be used as a basis for instrument zero and span
adjustments, calibration updates or adjustment of ambient data.
Whenever there is an indication of possible instrument drift or
malfunction, a full zero and calibration (Level 1) should be performed prior
to corrective action. For further information on zero, span and calibration
of air pollution monitors, refer to Section 2.0.9 of the US EPA's Quality
Assurance Handbook for Air Pollution Measurement Systems (Volume II).
Figure 9–9 shows how this option is integrated with the 43iQ components.
During normal operation, the pump draws zero air through the permeation
oven and out the instrument exhaust. When performing a single point span
check, the sample valve is energized, closing the sample valve, and the Perm
valve is energized, opening the perm valve. This directs the zero air/SO2 gas
mixture from the permeation oven gas through the heated hydrocarbon
kicker and into the optical bench.
Thermo Scientific
43iQ Instruction Manual
9-17
Optional Equipment
Internal Permeation Span Source
Figure 9–9. Internal Permeation Span Source Flow Diagram
9-18
43iQ Instruction Manual
Thermo Scientific
Optional Equipment
Internal Permeation Span Source
Permeation Tube
Installation
Use the following procedure to install the permeation tube.
CAUTION Allow oven to cool down prior to servicing. ▲
1. Lift oven cover up.
2. Lift and unlatch two oven cover handles.
Figure 9–10. Installing Glass Tube pt 1
3. Unfasten nut. Loosen thumb screw 2X full counter clock rotation.
4. Loosening the thumb screw deflates the o-ring making it easier to pull
the thermistor assembly out.
Thermo Scientific
43iQ Instruction Manual
9-19
Optional Equipment
Internal Permeation Span Source
5. Pull thermistor assembly upwards.
Figure 9–11. Installing Glass Tube pt 2
6. Push and gently twist tube dispersion glass upwards and seat into oring.
Note Make sure PTFE tube and thermistor is inside the dispersion glass.
Keep glass clean by using kimwipes or similar material to handle glass. ▲
9-20
43iQ Instruction Manual
Thermo Scientific
Optional Equipment
Internal Permeation Span Source
Figure 9–12. Installing Glass Tube pt 3
7. Insert thermistor assembly into oven tube. Make sure it bottoms out.
8. Tighten thumb screw to expand o-ring for complete seal.
9. Fasten nut.
Thermo Scientific
43iQ Instruction Manual
9-21
Optional Equipment
Internal Permeation Span Source
Figure 9–13. Installing Glass Tube pt 4
10. Latch oven cover handles.
Computation of
Concentrations
The computation of SO2 output level is shown in the following
information. Note that is assumed that all devices are properly calibrated
and that all flows are corrected to 25 °C and 1 atm.
Permeation Tube:
Output (ppm) =
(R) (K)
Qo
Where:
R = permeation rate in ng/min
Q0 = flow rate of gas (scc/min) into the charcoal scrubber during span mode
K = constant for the specific permeant = 24.45 / MW
MW = molecular weight
9-22
43iQ Instruction Manual
Thermo Scientific
Optional Equipment
PTFE Particulate Filter
PTFE Particulate
Filter
Thermo Scientific
A 5-10 micron pore size, two-inch diameter PTFE element is available for
the 43iQ. This filter should be installed just prior to the SAMPLE bulkhead.
When using a filter, all calibrations and span checks must be performed
through the filter.
43iQ Instruction Manual
9-23
Appendix A
Safety, Warranty, and WEEE
Safety
Safety and
Equipment Damage
Alerts
Review the following information carefully before using the instrument.
This manual provides specific information on how to operate the
instrument, however if the instrument is used in a manner not specified by
the manufacturer, the protection provided by the equipment may be
impaired.
This manual contains important information to alert you to potential safety
hazards and risks of equipment damage. Refer to the following types of
alerts you may see in this manual.
Safety and Equipment Damage Alert Descriptions
Description
Alert
A hazard is present that will result in death or serious
personal injury if the warning is ignored. ▲
A hazard is present or an unsafe practice can result in
serious personal injury if the warning is ignored. ▲
The hazard or unsafe practice could result in minor to
moderate personal injury if the warning is ignored. ▲
Equipment Damage
The hazard or unsafe practice could result in property
damage if the warning is ignored. ▲
Safety and Equipment Damage Alerts in this Manual
Description
Alert
If the equipment is operated in a manner not specified by
the manufacturer, the protection provided by the equipment
may be impaired. ▲
The service procedures in this manual are restricted to
qualified service personnel only. ▲
Equipment Damage
Do not attempt to lift the analyzer by the cover or other
external fittings. ▲
This adjustment should only be performed by an instrument
service technician. ▲
Thermo Scientific
43iQ Instruction Manual
A-1
Safety, Warranty, and WEEE
Warranty
Warranty
Seller warrants that the Products will operate or perform substantially in
conformance with Seller's published specifications and be free from defects
in material and workmanship, when subjected to normal, proper and
intended usage by properly trained personnel, for the period of time set
forth in the product documentation, published specifications or package
inserts. If a period of time is not specified in Seller’s product
documentation, published specifications or package inserts, the warranty
period shall be one (1) year from the date of shipment to Buyer for
equipment and ninety (90) days for all other products (the "Warranty
Period"). Seller agrees during the Warranty Period, to repair or replace, at
Seller's option, defective Products so as to cause the same to operate in
substantial conformance with said published specifications; provided that
(a) Buyer shall promptly notify Seller in writing upon the discovery of any
defect, which notice shall include the product model and serial number (if
applicable) and details of the warranty claim; (b) after Seller’s review, Seller
will provide Buyer with service data and/or a Return Material
Authorization (“RMA”), which may include biohazard decontamination
procedures and other product-specific handling instructions; and (c) then,
if applicable, Buyer may return the defective Products to Seller with all
costs prepaid by Buyer. Replacement parts may be new or refurbished, at
the election of Seller. All replaced parts shall become the property of Seller.
Shipment to Buyer of repaired or replacement Products shall be made in
accordance with the Delivery provisions of the Seller’s Terms and
Conditions of Sale. Consumables, including but not limited to lamps,
fuses, batteries, bulbs and other such expendable items, are expressly
excluded from the warranty under this warranty.
Notwithstanding the foregoing, Products supplied by Seller that are
obtained by Seller from an original manufacturer or third party supplier are
not warranted by Seller, but Seller agrees to assign to Buyer any warranty
rights in such Product that Seller may have from the original manufacturer
or third party supplier, to the extent such assignment is allowed by such
original manufacturer or third party supplier.
In no event shall Seller have any obligation to make repairs, replacements
or corrections required, in whole or in part, as the result of (i) normal wear
and tear, (ii) accident, disaster or event of force majeure, (iii) misuse, fault
or negligence of or by Buyer, (iv) use of the Products in a manner for which
they were not designed, (v) causes external to the Products such as, but not
limited to, power failure or electrical power surges, (vi) improper storage
and handling of the Products or (vii) use of the Products in combination
with equipment or software not supplied by Seller. If Seller determines
that Products for which Buyer has requested warranty services are not
covered by the warranty hereunder, Buyer shall pay or reimburse Seller for
all costs of investigating and responding to such request at Seller's then
A-2
43iQ Instruction Manual
Thermo Scientific
Safety, Warranty, and WEEE
Warranty
prevailing time and materials rates. If Seller provides repair services or
replacement parts that are not covered by the warranty provided in this
warranty, Buyer shall pay Seller therefor at Seller's then prevailing time and
materials rates. ANY INSTALLATION, MAINTENANCE, REPAIR,
SERVICE, RELOCATION OR ALTERATION TO OR OF, OR
OTHER TAMPERING WITH, THE PRODUCTS PERFORMED BY
ANY PERSON OR ENTITY OTHER THAN SELLER WITHOUT
SELLER'S PRIOR WRITTEN APPROVAL, OR ANY USE OF
REPLACEMENT PARTS NOT SUPPLIED BY SELLER, SHALL
IMMEDIATELY VOID AND CANCEL ALL WARRANTIES WITH
RESPECT TO THE AFFECTED PRODUCTS.
THE OBLIGATIONS CREATED BY THIS WARRANTY
STATEMENT TO REPAIR OR REPLACE A DEFECTIVE PRODUCT
SHALL BE THE SOLE REMEDY OF BUYER IN THE EVENT OF A
DEFECTIVE PRODUCT. EXCEPT AS EXPRESSLY PROVIDED IN
THIS WARRANTY STATEMENT, SELLER DISCLAIMS ALL
OTHER WARRANTIES, WHETHER EXPRESS OR IMPLIED, ORAL
OR WRITTEN, WITH RESPECT TO THE PRODUCTS,
INCLUDING WITHOUT LIMITATION ALL IMPLIED
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY
PARTICULAR PURPOSE. SELLER DOES NOT WARRANT THAT
THE PRODUCTS ARE ERROR-FREE OR WILL ACCOMPLISH
ANY PARTICULAR RESULT.
Thermo Scientific
43iQ Instruction Manual
A-3
Safety, Warranty, and WEEE
WEEE Compliance
WEEE
Compliance
This product is required to comply with the European Union’s Waste
Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is
marked with the following symbol:
Thermo Fisher Scientific has contracted with one or more
recycling/disposal companies in each EU Member State, and this product
should be disposed of or recycled through them. Further information on
Thermo Fisher Scientific’s compliance with these Directives, the recyclers
in your country, and information on Thermo Fisher Scientific products
which may assist the detection of substances subject to the RoHS Directive
are available at: www.thermoscientific.com/WEEERoHS.
WEEE Symbol
The following symbol and description identify the WEEE marking used on
the instrument and in the associated documentation.
Symbol
Description
Marking of electrical and electronic equipment which applies to electrical and
electronic equipment falling under the Directive 2002/96/EC (WEEE) and the
equipment that has been put on the market after 13 August 2005. ▲
A-4
43iQ Instruction Manual
Thermo Scientific
Appendix B
Quick Reference
List of Figures
Thermo Scientific
Figure 1–1. 43iQ Front ........................................................................................ 1-2
Figure 1–2. 43iQ Flow Schematic ....................................................................... 1-4
Figure 1–3. 43iQ Flow Schematic with Zero Span ............................................. 1-4
Figure 1–4. Bench Mount Assembly (dimensions in inches [mm]) .................... 1-7
Figure 1–5. Rack Mount Assembly (dimensions in inches [mm]) ...................... 1-8
Figure 1–6. Rack Mount Requirements .............................................................. 1-9
Figure 1–7. Rack Requirements Part 2 ............................................................... 1-9
Figure 2–1. Removing the Cover......................................................................... 2-2
Figure 2–2. Installing Feet .................................................................................. 2-3
Figure 2–3. Removing the Front Panel ............................................................... 2-4
Figure 2–4. Installing Ears and Handles ............................................................. 2-5
Figure 2–5. 43iQ Rear Panel ............................................................................... 2-7
Figure 2–6. Atmospheric Dump Bypass Plumbing ............................................. 2-7
Figure 2–7. Front Panel and Touchscreen Display ............................................. 2-8
Figure 4–1. Cylinder Gas Dilution System.......................................................... 4-3
Figure 4–2. Permeation Tube System ................................................................ 4-5
Figure 5–1. Single Stage Pump .......................................................................... 5-2
Figure 5–2. Pump Rebuilding .............................................................................. 5-4
Figure 7–1. Properly Grounded Antistatic Wrist Strap ...................................... 7-2
Figure 7–2. 43iQ Component Layout Top View .................................................. 7-4
Figure 7–3. 43iQ Component Layout Side View................................................. 7-4
Figure 7–4. Replacing the Fuses......................................................................... 7-5
Figure 7–5. Start with Top Right Corner of Fan Cover ....................................... 7-6
Figure 7–6. Removing the Fan Cover .................................................................. 7-6
Figure 7–7. Unplugging the Fan Cable ............................................................... 7-7
Figure 7–8. Replacing the Fan ............................................................................ 7-8
Figure 7–9. Unplugging the Fan Cable ............................................................... 7-9
Figure 7–10. Unplugging the DMC Cable......................................................... 7-10
Figure 7–11. Unfasten Hardware Front for Measurement Side Removal ....... 7-10
Figure 7–12. Unfasten Hardware Rear for Measurement Side Removal ........ 7-11
Figure 7–13. Replacing the LCD Module .......................................................... 7-12
Figure 7–14. Remove Electrical Cables from LCD ............................................ 7-13
43iQ Instruction Manual
B-1
Quick Reference
List of Figures
Figure 7–15. I/O Replacement, Arm ................................................................. 7-14
Figure 7–16. I/O Replacement, Remove Board ................................................ 7-14
Figure 7–17. I/O Replacement, Install .............................................................. 7-15
Figure 7–18. I/O Replacement, Close Arm Alignment ..................................... 7-15
Figure 7–19. Replacing the Peripherals Support Board ................................... 7-16
Figure 7–20. Flow Pressure Board, Disconnect DMC Cable ............................ 7-17
Figure 7–21. Flow Pressure Board, Screws ...................................................... 7-18
Figure 7–22. Remove Pump, Disconnect and Unfasten ................................... 7-20
Figure 7–23. Pump Removal, Keyway............................................................... 7-21
Figure 7–24. Pump replacing, Unfasten Screws .............................................. 7-21
Figure 7–25. Remove Capillary, Disconnect and Unfasten .............................. 7-22
Figure 7–26. Capillary Plate, Keyway ............................................................... 7-23
Figure 7–27. Capillary and O-Ring Replace ...................................................... 7-23
Figure 7–28. Removing Power Supply .............................................................. 7-25
Figure 7–29. Replacing Power Supply .............................................................. 7-25
Figure 7–30. Unplug and Unfasten Step POL Board......................................... 7-26
Figure 7–31. Clear Partition Keyway Step POL Board ...................................... 7-27
Figure 7–32. Step POL Board 1 Switch Settings .............................................. 7-28
Figure 7–33. Optional Step POL Board 2 Switch Settings ............................... 7-28
Figure 7–34. Removing the DMC Optical Bench Removal from Instrument .... 7-30
Figure 7–35. Removing the Optical Bench Assembly....................................... 7-31
Figure 7–36. Removing the Optical Bench Assembly pt 2 ............................... 7-32
Figure 7–37. Removing the PMT Assembly ...................................................... 7-33
Figure 7–38. Removing the PMT Cover ............................................................ 7-34
Figure 7–39. Replacing the PMT ....................................................................... 7-34
Figure 7–40. Removing and Replacing the Flasher Pack.................................. 7-35
Figure 7–41. Replacing the Flasher Pack Lamp ................................................ 7-36
Figure 7–42. Replacing the Photo Lamp Detector Board pt 1 .......................... 7-37
Figure 7–43. Replacing the Photo Lamp Detector board pt 2 .......................... 7-38
Figure 7–44. Removing and Replacing the Kicker Assembly ........................... 7-39
Figure 7–45. Replacing the Manifold pt 1 ........................................................ 7-40
Figure 7–46. Replacing the Manifold pt 2 ........................................................ 7-41
Figure 7–47. Replacing the Manifold pt 3 ........................................................ 7-41
Figure 7–48. Removing and Replacing the Solenoid Valve pt 1 ...................... 7-42
Figure 7–49. Removing and Replacing the Solenoid Valve pt 2 ...................... 7-43
Figure 7–50. Perm Oven pt 1 ............................................................................. 7-44
Figure 7–51. Finger Push and Release Tubing ................................................. 7-45
Figure 7–52. Perm Oven Replacement pt 2 ...................................................... 7-46
Figure 7–53. Replacing the Permeation Oven Board........................................ 7-47
Figure 8–1. Common System Interconnect Diagram .......................................... 8-5
B-2
43iQ Instruction Manual
Thermo Scientific
Quick Reference
List of Tables
Figure 8–2. 43iQ System Interconnect Diagram ................................................ 8-6
Figure 9–1. I/O Expansion Replacement Boards ................................................ 9-1
Figure 9–2. RS-232/RS-485 Port......................................................................... 9-2
Figure 9–3. RS-485 External Accessory Port ...................................................... 9-3
Figure 9–4. Analog Voltage Inputs ..................................................................... 9-4
Figure 9–5. Analog Voltage and Current ............................................................ 9-5
Figure 9–6. Digital Inputs ................................................................................. 9-11
Figure 9–7. Digital Relay Switches .................................................................. 9-13
Figure 9–8. Valve Driver Outputs ..................................................................... 9-15
Figure 9–9. Internal Permeation Span Source Flow Diagram.......................... 9-18
Figure 9–10. Installing Glass Tube pt 1 ............................................................ 9-19
Figure 9–11. Installing Glass Tube pt 2 ............................................................ 9-20
Figure 9–12. Installing Glass Tube pt 3 ............................................................ 9-21
Figure 9–13. Installing Glass Tube pt 4 ............................................................ 9-22
List of Tables
Thermo Scientific
Table 1–1. 43iQ Specifications ........................................................................... 1-5
Table 1–2. 43iQ Optional Permeation Oven Specifications ............................... 1-6
Table 3–1. Data Logging Variables .................................................................. 3-30
Table 3–2. Streaming Data Variables .............................................................. 3-33
Table 6–1. 43iQ Troubleshooting Guide ............................................................. 6-1
Table 9–1. RS-232/RS-485 Port Terminal Assignment ...................................... 9-2
Table 9–2. RS-485 External Accessory Port Terminal Assignment ................... 9-3
Table 9–3. Analog Voltage Inputs Assignment .................................................. 9-4
Table 9–4. Analog Voltage and Current Assignment ......................................... 9-6
Table 9–5. Digital Inputs Terminal Assignment ............................................... 9-12
Table 9–6. Digital Relay Switch Assignment ................................................... 9-14
Table 9–7. Valve Driver Outputs Assignment .................................................. 9-16
43iQ Instruction Manual
B-3
Appendix C
GNU Lesser General Public License
GNU Lesser
General Public
License
Version 2.1, February 1999
Copyright (C) 1991, 1999 Free Software Foundation, Inc.
51 Franklin Street, Fifth Floor, Boston, MA 02110-1301 USA
Everyone is permitted to copy and distribute verbatim copies of this license
document, but changing it is not allowed.
BECAUSE THE LIBRARY IS LICENSED FREE OF CHARGE,
THERE IS NO WARRANTY FOR THE LIBRARY, TO THE EXTENT
PERMITTED BY APPLICABLE LAW.
The Open Source code used in the development of this product will be
provided for a period of 3-years from date of purchase to the original
purchaser provided they supply Thermo Fisher Scientific with the
applicable serial number for the product the Open Source code is being
requested for.
[This is the first released version of the Lesser GPL. It also counts as the
successor of the GNU Library Public License, version 2, hence the version
number 2.1.]
Preamble
The licenses for most software are designed to take away your freedom to
share and change it. By contrast, the GNU General Public Licenses are
intended to guarantee your freedom to share and change free software--to
make sure the software is free for all its users.
This license, the Lesser General Public License, applies to some specially
designated software packages--typically libraries--of the Free Software
Foundation and other authors who decide to use it. You can use it too, but
we suggest you first think carefully about whether this license or the
ordinary General Public License is the better strategy to use in any
particular case, based on the explanations below.
When we speak of free software, we are referring to freedom of use, not
price. Our General Public Licenses are designed to make sure that you
have the freedom to distribute copies of free software (and charge for this
service if you wish); that you receive source code or can get it if you want it;
Thermo Scientific
43iQ Instruction Manual
C-1
GNU Lesser General Public License
GNU Lesser General Public License
that you can change the software and use pieces of it in new free programs;
and that you are informed that you can do these things.
To protect your rights, we need to make restrictions that forbid distributors
to deny you these rights or to ask you to surrender these rights. These
restrictions translate to certain responsibilities for you if you distribute
copies of the library or if you modify it.
For example, if you distribute copies of the library, whether gratis or for a
fee, you must give the recipients all the rights that we gave you. You must
make sure that they, too, receive or can get the source code. If you link
other code with the library, you must provide complete object files to the
recipients, so that they can relink them with the library after making
changes to the library and recompiling it. And you must show them these
terms so they know their rights.
We protect your rights with a two-step method: (1) we copyright the
library, and (2) we offer you this license, which gives you legal permission
to copy, distribute and/or modify the library.
To protect each distributor, we want to make it very clear that there is no
warranty for the free library. Also, if the library is modified by someone
else and passed on, the recipients should know that what they have is not
the original version, so that the original author's reputation will not be
affected by problems that might be introduced by others.
Finally, software patents pose a constant threat to the existence of any free
program. We wish to make sure that a company cannot effectively restrict
the users of a free program by obtaining a restrictive license from a patent
holder. Therefore, we insist that any patent license obtained for a version
of the library must be consistent with the full freedom of use specified in
this license.
Most GNU software, including some libraries, is covered by the ordinary
GNU General Public License. This license, the GNU Lesser General
Public License, applies to certain designated libraries, and is quite different
from the ordinary General Public License. We use this license for certain
libraries in order to permit linking those libraries into non-free programs.
When a program is linked with a library, whether statically or using a
shared library, the combination of the two is legally speaking a combined
work, a derivative of the original library. The ordinary General Public
License therefore permits such linking only if the entire combination fits its
criteria of freedom. The Lesser General Public License permits more lax
criteria for linking other code with the library.
We call this license the "Lesser" General Public License because it does Less
to protect the user's freedom than the ordinary General Public License. It
also provides other free software developers Less of an advantage over
competing non-free programs. These disadvantages are the reason we use
C-2
43iQ Instruction Manual
Thermo Scientific
GNU Lesser General Public License
GNU Lesser General Public License
the ordinary General Public License for many libraries. However, the
Lesser license provides advantages in certain special circumstances.
For example, on rare occasions, there may be a special need to encourage
the widest possible use of a certain library, so that it becomes a de-facto
standard. To achieve this, non-free programs must be allowed to use the
library. A more frequent case is that a free library does the same job as
widely used non-free libraries. In this case, there is little to gain by limiting
the free library to free software only, so we use the Lesser General Public
License.
In other cases, permission to use a particular library in non-free programs
enables a greater number of people to use a large body of free software. For
example, permission to use the GNU C Library in non-free programs
enables many more people to use the whole GNU operating system, as well
as its variant, the GNU/Linux operating system.
Although the Lesser General Public License is Less protective of the users'
freedom, it does ensure that the user of a program that is linked with the
Library has the freedom and the wherewithal to run that program using a
modified version of the Library.
The precise terms and conditions for copying, distribution and
modification follow. Pay close attention to the difference between a "work
based on the library" and a "work that uses the library". The former
contains code derived from the library, whereas the latter must be
combined with the library in order to run.
GNU LESSER GENERAL PUBLIC LICENSE TERMS AND
CONDITIONS FOR COPYING, DISTRIBUTION AND
MODIFICATION
0. This License Agreement applies to any software library or other program
which contains a notice placed by the copyright holder or other authorized
party saying it may be distributed under the terms of this Lesser General
Public License (also called "this License"). Each licensee is addressed as
"you".
A "library" means a collection of software functions and/or data prepared
so as to be conveniently linked with application programs (which use some
of those functions and data) to form executables.
The "Library", below, refers to any such software library or work which has
been distributed under these terms. A "work based on the Library" means
either the Library or any derivative work under copyright law: that is to say,
a work containing the Library or a portion of it, either verbatim or with
modifications and/or translated straightforwardly into another language.
(Hereinafter, translation is included without limitation in the term
"modification".)
Thermo Scientific
43iQ Instruction Manual
C-3
GNU Lesser General Public License
GNU Lesser General Public License
"Source code" for a work means the preferred form of the work for making
modifications to it. For a library, complete source code means all the
source code for all modules it contains, plus any associated interface
definition files, plus the scripts used to control compilation and installation
of the library.
Activities other than copying, distribution and modification are not
covered by this License; they are outside its scope. The act of running a
program using the Library is not restricted, and output from such a
program is covered only if its contents constitute a work based on the
Library (independent of the use of the Library in a tool for writing it).
Whether that is true depends on what the Library does and what the
program that uses the Library does.
1. You may copy and distribute verbatim copies of the Library's complete
source code as you receive it, in any medium, provided that you
conspicuously and appropriately publish on each copy an appropriate
copyright notice and disclaimer of warranty; keep intact all the notices that
refer to this License and to the absence of any warranty; and distribute a
copy of this License along with the Library.
You may charge a fee for the physical act of transferring a copy, and you
may at your option offer warranty protection in exchange for a fee.
2. You may modify your copy or copies of the Library or any portion of it,
thus forming a work based on the Library, and copy and distribute such
modifications or work under the terms of Section 1 above, provided that
you also meet all of these conditions:
a) The modified work must itself be a software library.
b) You must cause the files modified to carry prominent notices
that you changed the files and the date of any change.
stating
c) You must cause the whole of the work to be licensed at no charge to all
third parties under the terms of this License.
d) If a facility in the modified Library refers to a function or a table of data
to be supplied by an application program that uses the facility, other than
as an argument passed when the facility is invoked, then you must make a
good faith effort to ensure that, in the event an application does not supply
such function or table, the facility still operates, and performs whatever part
of its purpose remains meaningful.
(For example, a function in a library to compute square roots has a purpose
that is entirely well-defined independent of the application. Therefore,
Subsection 2d requires that any application-supplied function or table used
by this function must be optional: if the application does not supply it, the
square root function must still compute square roots.)
C-4
43iQ Instruction Manual
Thermo Scientific
GNU Lesser General Public License
GNU Lesser General Public License
These requirements apply to the modified work as a whole. If identifiable
sections of that work are not derived from the Library, and can be
reasonably considered independent and separate works in themselves, then
this License, and its terms, do not apply to those sections when you
distribute them as separate works. But when you distribute the same
sections as part of a whole which is a work based on the Library, the
distribution of the whole must be on the terms of this License, whose
permissions for other licensees extend to the entire whole, and thus to each
and every part regardless of who wrote it.
Thus, it is not the intent of this section to claim rights or contest your
rights to work written entirely by you; rather, the intent is to exercise the
right to control the distribution of derivative or collective works based on
the Library.
In addition, mere aggregation of another work not based on the Library
with the Library (or with a work based on the Library) on a volume of a
storage or distribution medium does not bring the other work under the
scope of this License.
3. You may opt to apply the terms of the ordinary GNU General Public
License instead of this License to a given copy of the Library. To do this,
you must alter all the notices that refer to this License, so that they refer to
the ordinary GNU General Public License, version 2, instead of to this
License. (If a newer version than version 2 of the ordinary GNU General
Public License has appeared, then you can specify that version instead if
you wish.) Do not make any other change in these notices.
Once this change is made in a given copy, it is irreversible for that copy, so
the ordinary GNU General Public License applies to all subsequent copies
and derivative works made from that copy. This option is useful when you
wish to copy part of the code of the Library into a program that is not a
library.
4. You may copy and distribute the Library (or a portion or derivative of it,
under Section 2) in object code or executable form under the terms of
Sections 1 and 2 above provided that you accompany it with the complete
corresponding machine-readable source code, which must be distributed
under the terms of Sections 1 and 2 above on a medium customarily used
for software interchange.
If distribution of object code is made by offering access to copy from a
designated place, then offering equivalent access to copy the source code
from the same place satisfies the requirement to distribute the source code,
even though third parties are not compelled to copy the source along with
the object code.
5. A program that contains no derivative of any portion of the Library, but
is designed to work with the Library by being compiled or linked with it, is
Thermo Scientific
43iQ Instruction Manual
C-5
GNU Lesser General Public License
GNU Lesser General Public License
called a "work that uses the Library". Such a work, in isolation, is not a
derivative work of the Library, and therefore falls outside the scope of this
License.
However, linking a "work that uses the Library" with the Library creates an
executable that is a derivative of the Library (because it contains portions of
the Library), rather than a "work that uses the library". The executable is
therefore covered by this License. Section 6 states terms for distribution of
such executables.
When a "work that uses the Library" uses material from a header file that is
part of the Library, the object code for the work may be a derivative work
of the Library even though the source code is not. Whether this is true is
especially significant if the work can be linked without the Library, or if the
work is itself a library. The threshold for this to be true is not precisely
defined by law.
If such an object file uses only numerical parameters, data structure layouts
and accessors, and small macros and small inline functions (ten lines or less
in length), then the use of the object file is unrestricted, regardless of
whether it is legally a derivative work. (Executables containing this object
code plus portions of the Library will still fall under Section 6.)
Otherwise, if the work is a derivative of the Library, you may distribute the
object code for the work under the terms of Section 6. Any executables
containing that work also fall under Section 6, whether or not they are
linked directly with the Library itself.
6. As an exception to the Sections above, you may also combine or link a
"work that uses the Library" with the Library to produce a work containing
portions of the Library, and distribute that work under terms of your
choice, provided that the terms permit modification of the work for the
customer's own use and reverse engineering for debugging such
modifications.
You must give prominent notice with each copy of the work that the
Library is used in it and that the Library and its use are covered by this
License. You must supply a copy of this License. If the work during
execution displays copyright notices, you must include the copyright notice
for the Library among them, as well as a reference directing the user to the
copy of this License. Also, you must do one of these things:
a) Accompany the work with the complete corresponding machinereadable source code for the Library including whatever changes were used
in the work (which must be distributed under Sections 1 and 2 above);
and, if the work is an executable linked with the Library, with the complete
machine-readable "work that uses the Library", as object code and/or
source code, so that the user can modify the Library and then relink to
produce a modified executable containing the modified Library. (It is
C-6
43iQ Instruction Manual
Thermo Scientific
GNU Lesser General Public License
GNU Lesser General Public License
understood that the user who changes the contents of definitions files in
the Library will not necessarily be able to recompile the application to use
the modified definitions.)
b) Use a suitable shared library mechanism for linking with the Library. A
suitable mechanism is one that (1) uses at run time a copy of the library
already present on the user's computer system, rather than copying library
functions into the executable, and (2) will operate properly with a modified
version of the library, if the user installs one, as long as the modified version
is interface-compatible with the version that the work was made with.
c) Accompany the work with a written offer, valid for at least three years, to
give the same user the materials specified in Subsection 6a, above, for a
charge no more than the cost of performing this distribution.
d) If distribution of the work is made by offering access to copy from a
designated place, offer equivalent access to copy the above specified
materials from the same place.
e) Verify that the user has already received a copy of these materials or that
you have already sent this user a copy.
For an executable, the required form of the "work that uses the Library"
must include any data and utility programs needed for reproducing the
executable from it. However, as a special exception, the materials to be
distributed need not include anything that is normally distributed (in either
source or binary form) with the major components (compiler, kernel, and
so on) of the operating system on which the executable runs, unless that
component itself accompanies the executable.
It may happen that this requirement contradicts the license restrictions of
other proprietary libraries that do not normally accompany the operating
system. Such a contradiction means you cannot use both them and the
Library together in an executable that you distribute.
7. You may place library facilities that are a work based on the Library sideby-side in a single library together with other library facilities not covered
by this License, and distribute such a combined library, provided that the
separate distribution of the work based on the Library and of the other
library facilities is otherwise permitted, and provided that you do these two
things:
a) Accompany the combined library with a copy of the same work based on
the Library, uncombined with any other library facilities. This must be
distributed under the terms of the Sections above.
b) Give prominent notice with the combined library of the fact that part of
it is a work based on the Library, and explaining where to find the
accompanying uncombined form of the same work.
Thermo Scientific
43iQ Instruction Manual
C-7
GNU Lesser General Public License
GNU Lesser General Public License
8. You may not copy, modify, sublicense, link with, or distribute the
Library except as expressly provided under this License. Any attempt
otherwise to copy, modify, sublicense, link with, or distribute the Library is
void, and will automatically terminate your rights under this License.
However, parties who have received copies, or rights, from you under this
License will not have their licenses terminated so long as such parties
remain in full compliance.
9. You are not required to accept this License, since you have not signed it.
However, nothing else grants you permission to modify or distribute the
Library or its derivative works. These actions are prohibited by law if you
do not accept this License. Therefore, by modifying or distributing the
Library (or any work based on the Library), you indicate your acceptance of
this License to do so, and all its terms and conditions for copying,
distributing or modifying the Library or works based on it.
10. Each time you redistribute the Library (or any work based on the
Library), the recipient automatically receives a license from the original
licensor to copy, distribute, link with or modify the Library subject to these
terms and conditions. You may not impose any further restrictions on the
recipients' exercise of the rights granted herein. You are not responsible for
enforcing compliance by third parties with this License.
11. If, as a consequence of a court judgment or allegation of patent
infringement or for any other reason (not limited to patent issues),
conditions are imposed on you (whether by court order, agreement or
otherwise) that contradict the conditions of this License, they do not excuse
you from the conditions of this License. If you cannot distribute so as to
satisfy simultaneously your obligations under this License and any other
pertinent obligations, then as a consequence you may not distribute the
Library at all. For example, if a patent license would not permit royaltyfree redistribution of the Library by all those who receive copies directly or
indirectly through you, then the only way you could satisfy both it and this
License would be to refrain entirely from distribution of the Library.
If any portion of this section is held invalid or unenforceable under any
particular circumstance, the balance of the section is intended to apply, and
the section as a whole is intended to apply in other circumstances.
It is not the purpose of this section to induce you to infringe any patents or
other property right claims or to contest validity of any such claims; this
section has the sole purpose of protecting the integrity of the free software
distribution system which is implemented by public license practices.
Many people have made generous contributions to the wide range of
software distributed through that system in reliance on consistent
application of that system; it is up to the author/donor to decide if he or
she is willing to distribute software through any other system and a licensee
cannot impose that choice.
C-8
43iQ Instruction Manual
Thermo Scientific
GNU Lesser General Public License
GNU Lesser General Public License
This section is intended to make thoroughly clear what is believed to be a
consequence of the rest of this License.
12. If the distribution and/or use of the Library is restricted in certain
countries either by patents or by copyrighted interfaces, the original
copyright holder who places the Library under this License may add an
explicit geographical distribution limitation excluding those countries, so
that distribution is permitted only in or among countries not thus
excluded. In such case, this License incorporates the limitation as if written
in the body of this License.
13. The Free Software Foundation may publish revised and/or new
versions of the Lesser General Public License from time to time. Such new
versions will be similar in spirit to the present version, but may differ in
detail to address new problems or concerns.
Each version is given a distinguishing version number. If the Library
specifies a version number of this License which applies to it and "any later
version", you have the option of following the terms and conditions either
of that version or of any later version published by the Free Software
Foundation. If the Library does not specify a license version number, you
may choose any version ever published by the Free Software Foundation.
14. If you wish to incorporate parts of the Library into other free programs
whose distribution conditions are incompatible with these, write to the
author to ask for permission. For software which is copyrighted by the
Free Software Foundation, write to the Free Software Foundation; we
sometimes make exceptions for this. Our decision will be guided by the
two goals of preserving the free status of all derivatives of our free software
and of promoting the sharing and reuse of software generally.
NO WARRANTY
15. BECAUSE THE LIBRARY IS LICENSED FREE OF CHARGE,
THERE IS NO WARRANTY FOR THE LIBRARY, TO THE EXTENT
PERMITTED BY APPLICABLE LAW. EXCEPT WHEN OTHERWISE
STATED IN WRITING THE COPYRIGHT HOLDERS AND/OR
OTHER PARTIES PROVIDE THE LIBRARY "AS IS" WITHOUT
WARRANTY OF ANY KIND, EITHER EXPRESSED OR IMPLIED,
INCLUDING, BUT NOT LIMITED TO, THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE. THE ENTIRE RISK AS TO THE
QUALITY AND PERFORMANCE OF THE LIBRARY IS WITH YOU.
SHOULD THE LIBRARY PROVE DEFECTIVE, YOU ASSUME THE
COST OF ALL NECESSARY SERVICING, REPAIR OR
CORRECTION.
16. IN NO EVENT UNLESS REQUIRED BY APPLICABLE LAW OR
AGREED TO IN WRITING WILL ANY COPYRIGHT HOLDER, OR
Thermo Scientific
43iQ Instruction Manual
C-9
GNU Lesser General Public License
GNU Lesser General Public License
ANY OTHER PARTY WHO MAY MODIFY AND/OR
REDISTRIBUTE THE LIBRARY AS PERMITTED ABOVE, BE
LIABLE TO YOU FOR DAMAGES, INCLUDING ANY GENERAL,
SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES
ARISING OUT OF THE USE OR INABILITY TO USE THE
LIBRARY (INCLUDING BUT NOT LIMITED TO LOSS OF DATA
OR DATA BEING RENDERED INACCURATE OR LOSSES
SUSTAINED BY YOU OR THIRD PARTIES OR A FAILURE OF
THE LIBRARY TO OPERATE WITH ANY OTHER SOFTWARE),
EVEN IF SUCH HOLDER OR OTHER PARTY HAS BEEN
ADVISED OF THE POSSIBILITY OF SUCH DAMAGES.
END OF TERMS AND CONDITIONS
How to Apply These Terms to Your New Libraries
If you develop a new library, and you want it to be of the greatest possible
use to the public, we recommend making it free software that everyone can
redistribute and change. You can do so by permitting redistribution under
these terms (or, alternatively, under the terms of the ordinary General
Public License).
To apply these terms, attach the following notices to the library. It is safest
to attach them to the start of each source file to most effectively convey the
exclusion of warranty; and each file should have at least the "copyright" line
and a pointer to where the full notice is found.
Copyright (C)
This library is free software; you can redistribute it and/or modify it under
the terms of the GNU Lesser General Public License as published by the
Free Software Foundation; either version 2.1 of the License, or (at your
option) any later version.
This library is distributed in the hope that it will be useful, but
WITHOUT ANY WARRANTY; without even the implied warranty of
MERCHANTABILITY or FITNESS FOR A PARTICULAR PURPOSE.
See the GNU Lesser General Public License for more details.
You should have received a copy of the GNU Lesser General Public License
along with this library; if not, write to the Free Software Foundation, Inc.,
51 Franklin Street, Fifth Floor, Boston, MA 02110-1301 USA
Also add information on how to contact you by electronic and paper mail.
You should also get your employer (if you work as a programmer) or your
school, if any, to sign a "copyright disclaimer" for the library, if necessary.
Here is a sample; alter the names:
C-10
43iQ Instruction Manual
Thermo Scientific
GNU Lesser General Public License
GNU Lesser General Public License
Yoyodyne, Inc., hereby disclaims all copyright interest in the library `Frob'
(a library for tweaking knobs) written by James Random Hacker.
, 1 April 1990
Ty Coon, President of Vice
That's all there is to it!
Nokia Qt LGPL Exception version 1.1
As an additional permission to the GNU Lesser General Public License
version 2.1, the object code form of a ""work that uses the Library"" may
incorporate material from a header file that is part of the Library. You may
distribute such object code under terms of your choice, provided that:
(i) the header files of the Library have not been modified; and
(ii) the incorporated material is limited to numerical parameters, data
structure layouts, accessors, macros, inline functions and templates; and
(iii) you comply with the terms of Section 6 of the GNU Lesser General
Public License version 2.1.
Moreover, you may apply this exception to a modified version of the
Library, provided that such modification does not involve copying material
from the Library into the modified Library's header files unless such
material is limited to (i) numerical parameters; (ii) data structure layouts;
(iii) accessors; and (iv) small macros, templates and inline functions of five
lines or less in length.
Furthermore, you are not required to apply this additional permission to a
modified version of the Library.
Thermo Scientific
43iQ Instruction Manual
C-11
USA
27 Forge Parkway
Franklin, MA 02038
Ph: (508) 520-0430
Fax: (508) 520-2800
India
C/327, TTC Industrial Area
MIDC Pawane
New Mumbai 400 705, India
Ph: +91 22 4157 8800
China
+Units 702-715, 7th Floor
Tower West, Yonghe
Beijing, China 100007
Ph: +86 10 84193588
Europe
Ion Path, Road Three,
Winsford, Cheshire CW73GA UK
Ph: +44 1606 548700
Fax: +44 1606 548711
orders.aqi@thermofisher.com
india@thermofisher.com
info.eid.china@thermofisher.com
sales.epm.uk@thermofisher.com
Find out more at thermofisher.com/43iQ
© 2017 Thermo Fisher Scientific Inc. All trademarks are the property of Thermo Fisher Scientific and its
subsidiaries unless otherwise specified. EPM_117568-00