Draeger / Drager Introduction to Gas Detection Systems
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
ST-6057-2004
Introduction to
Gas Detection Systems
ST-6028-2004
STL-1097-2008
What is Gaseous Matter?
Matter with a temperature above its boiling point is said to
be a gas. In respect to the environment of people (normal
conditions) any substance having a boiling point below 20 °C
at normal pressure is a gas. The most light-weight gas is
hydrogen (H2, 14-times lighter than air), the heaviest gas is
tungsten hexafluoride (WF6, approx. ten-times heavier than air).
Under normal conditions 1 cm3 gas contains about 30·1018
molecules, the average distance in between is only about
three nanometers. They swarm through the space by some
100 to 1000 meters per second, collide some billion times
per second with other molecules, so that between two
collisions they can only fly approx. 50 to 100 nanometers.
And with each collision they change their flight direction
and transfer energy to the collision partner.
This is an entirely random movement of molecules, which
macroscopically is measurable as gas temperature (average
kinetic energy of all molecules) and gas pressure (average
momentum of all molecules striking a surface) or extension
(volume). Therefore pressure, temperature and volume
mathematically are in a fix relationship, which in the ideal
case follow the so called ideal gas law:
- at constant pressure a gas volume changes
proportional to its temperature – e.g. it expands when
being heated
- at constant gas volume (gas in a closed vessel) the
gas pressure changes proportional to its temperature –
e.g. the inner pressure of a closed vessel rises when being
heated
- at constant temperature the gas pressure changes
proportional to its volume – e.g. the pressure rises when
the gas is compressed
Also, the extremely quick random movement of gas
molecules is the reason that gases mix up easily and never
ever will segregate from each other. The movement of
molecules into the direction of lower concentration (so-called
diffusion) is based on these molecular characteristics and
plays an essential role in the measuring principles of gas
sensors. Commonly diffusion processes are quicker the
faster the molecules move (the hotter the gas is) and the
lower the molar weight is (the lighter the gas is).
What is Gaseous Matter?
STL-1098-2008
Vapours? Why not Gases?
In opposite to gases – there might exist only approx. 200 to
300 – the proper term for the gaseous state of matter below
its boiling point is vapour. Vapour mostly exists in an
equilibrium with its liquid (or sometimes even solid) phase
and condenses or evaporates depending on its temperature.
This behaviour is well-known from water: Cooling down the
humid air during nighttime causes fog (condensation), while
the warmth of the morning sun dissolves the fog again
(evaporation).
In a closed vessel, above the surface of a liquid the maximum possible vapour concentration is always formed. This
vapour concentration depends on the liquid’s temperature.
In a microscopic view, vapour is generated by the random
movement of the liquid’s molecules and their ability that they
can overcome the surface tension and mix-up with the air
molecules above.
Each liquid has a certain characteristic vapour pressure
which only depends on the liquid’s temperature, and this
pressure is equal to the atmospheric pressure when its
temperature reaches the boiling point. The graph of this
correlation is known as the vapour pressure curve, enabling
us to determine the maximum possible vapour concentration
at a given temperature.
Vapour pressure curve of liquid n-Hexane
300
250
200
ST-1099-2008-e
150
100
50
0
-10
-5
0
5
10
15
20
25
30
35 ° C
Dividing the maximum possible vapour pressure by the
ambient pressure results in the so-called saturation
concentration in Vol%. E.g., for n-hexane at 20 °C (vapour
pressure 162 mbar) at an ambient pressure of 1000 mbar
the maximum possible n-hexane concentration is 16.2 Vol%.
Vapours? Why not Gases?
Our Atmosphere
While decreasing its specific weight continuously, our
atmosphere is extending far into the deep space. The sky’s
blue colour is caused by scattering of the sun light at the
air’s molecules (mainly nitrogen and oxygen), but at a height
of about 21 km the sky is black. Compressing the whole
atmosphere to constant pressure of roughly 1 bar its height
would be 8 km and the UV-absorbing stratospheric ozone
layer would be only 3 mm thick.
Typical composition of the earth’s atmosphere in ppm:
Gas
Main gases
N2 - Nitrogen
O2 - Oxygen
H2O - Water vapour
Ar - Argon
CO2 - Carbon dioxide
Trace gases
Ne - Neon
He - Helium
CH4 - Methane
Kr - Krypton
H2 - Hydrogen
N2O - Nitrous oxide
CO - Carbon monoxide
Xe - Xenon
O3 - Ozone
further trace gases
in total
dry
Composition
humid
780 840
209 450
0
9 340
340
768 543
206 152
15 748
9 193
335
18
5
1.8
1.1
0.5
0.3
0.09
0.09
0.07
3.05
1000 000
18
5
1.8
1.1
0.5
0.3
0.09
0.09
0.07
3.0
1000 000
1 Vol% = 10 000 ppm (= parts per million); assumption: rel. humidity 68 % RH at 20 °C
The mass of the earth’s atmosphere is about more than
5 000 trillion tons (5.235·1018 kg) pressing down on the
earth’s surface of 500 trillion square meters (0.507·1015 m2).
This is the reason why we have an atmospheric
pressure of 10 325 kg/m2, or, more familiar: 1 013 mbar. With
increasing height the atmospheric pressure decreases:
Altitude
-1000 m
- 500 m
0m
500 m
1000 m
1500 m
Pressure
1148 mbar
1078 mbar
1013 mbar
952 mbar
900 mbar
840 mbar
Altitude
2000 m
3000 m
4000 m
5000 m
6000 m
8000 m
Pressure
795 mbar
701 mbar
616 mbar
540 mbar
472 mbar
356 mbar
Since less molecules are present in a given volume at
lower atmospheric pressure, the signal of partial pressure
measuring gas detectors is always depending on the
atmospheric pressure.
Our Atmosphere
Oxygen
While nitrogen, with more than 78 Vol% the main gas of our
atmosphere, is completely inert (despite its excess it can not
even be used in this gaseous state by plants as a much
needed fertilizer), oxygen is very reactive and the base of our
breathing and existence, more over: the base of nearly every
being.
There is nearly 21 Vol% oxygen in our atmosphere. Oxygen
deficiency is life threatening and cannot be detected by our
nose. In general oxygen deficiency is caused by the release
of inert gases which displace oxygen. Since roughly one fifth
of the atmosphere is oxygen, the oxygen concentration is
only reduced by a fifth of the inert gas concentration.
If e.g. 10 Vol% of helium is released into the atmosphere the
oxygen concentration decreases by 2 Vol% while the nitrogen
concentration is reduced by 8 Vol%.
The industrial use of liquefied nitrogen (- 196 °C) in great
amounts may cause dangerous oxygen deficiency due to
leaks and sudden nitrogen evaporation.
Enhanced oxygen concentrations (e.g. more than 25
Vol%) cannot be sensed by people, but have severe consequences in respect to the flammability characteristics of
materials, and may even cause auto-ignition.
This is the reason why explosion protection measures are
only related to the atmospheric oxygen concentration.
When will it be dangerous?
Oxygen
concentration
in Vol%
below 17
11 to 14
8 to 11
6 to 8
less than 6
Oxygen
Symptoms
partial pressure
in mbar
below 170
Tendency of danger by
oxygen deficiency
110 to 140
Unperceived decrease of
physical and mental
capabilities
80 to 110
Possibility of sudden
unconsciousness without
warning after a certain time
of exposure
60 to 80
Loss of consciousness within
a few minutes, resuscitation
possible if instantly done
less than 60 Sudden loss of
consciousness
Oxygen
ST-4766-2004
Ex - Ox - Tox ... Danger!
Virtually, gases and vapours are always dangerous! If gases
do not exist in their familiar and respirable atmospheric
composition, safe breathing might already be affected.
Moreover: Any gas is potentially dangerous, being liquefied,
compressed or in normal state - essential is only their
concentration.
Basically there are three categories of risk
- Ex – Risk of explosion by flammable gases
- Ox – Oxygen
Risk of asphyxiation by oxygen displacement
Risk of increase of flammability by oxygen enrichment
- Tox – Risk of poisoning by toxic gases
Without auxiliary tools humans are not able to recognize
these dangers early enough to initiate appropriate counter
measures. With only some exceptions our nose has turned
out to be an extremely unreliable warning instrument.
For example, hydrogen sulfide in low concentrations is
sensible by its typical odor of rotten eggs, but lethal high
concentrations of hydrogen sulfide are not detectable by
our nose. Escaping into areas assumed to be non-dangerous
because of the missing smell has already caused a lot of
fatal accidents.
Even harmless gases such as argon, helium or nitrogen
may become dangerous when the vitally important oxygen is
displaced by a sudden release. Danger of suffocation!
Oxygen concentrations of less than 6 Vol% are known to be
lethal. Oxygen in excess (more than 21 Vol%) increases the
danger of flammability and might even cause auto-ignition of
flammable material. Not only can flammable gases and
vapours cause considerable plant damages by ignition, also
human life is compromised.
It is essential to detect Ex-Ox-Tox dangers reliably and to
protect human life, assets and environment by means of
appropriate measures.
- This is what gas detectors are required for
- This is what gas detection systems are required for.
Ex - Ox - Tox ... Danger!
Toxic Gases and Vapours
The toxicity of industrially utilized gases and vapours is
determined by means of laboratory experiments resulting in
the LC50 rate. From this and from further scientific and
occupational health investigations with even lower concentrations authorized committees deduce proposals for threshold
limit values (occupational exposure limits) that shall become
mandatory.
These threshold limit values are defined so that an employee
will not be harmed as long as he is not breathing in higher
gas concentrations as the stated threshold limit value during
his whole working life – however, this has to be ensured!
The concentrations of these limit values range over
seven magnitudes:
limit value
5000
1000
500
200
100
50
ppm
ppm
ppm
ppm
ppm
ppm
20 ppm
10 ppm
5 ppm
1 ppm
500 ppb
200 ppb
100 ppb
50 ppb
20 ppb
10 ppb
5 ppb
representative substance
with this limit value
Carbon dioxide
Propane, Butane
Acetone
Methyl ethyl ketone (MEK)
Butanol
n-Hexane, Toluene
Acetonitrile
Chlorobenzene
Diethyl amine
1.1.2.2-Tetrachloroethane
Chlorine
Methyl chloroformate
Chlorine dioxide
Decaborane
Phosgene
Methylisocyanate
MDI (Methyldiphenyldiisocyanate)
T+ very toxic
LC50 < 0.5 g/m3
Boron trichloride, Boron trifluoride, Bromine, Diborane, Fluorine,
Hydrogen arsenide, Hydrogen cyanide, Hydrogen fluoride,
Hydrogen phosphide, Hydrogen sulfide, Nitrogen dioxide,
Nitrogen monoxide, Ozone, Phosgene, Sulfur tetrafluoride,
Tungsten hexafluoride
T Toxic
LC50 = 0.5 ... 2.0 g/m3
Acetonitrile, Ammonia, Benzene, Carbon disulfide, Carbon
monoxide, Chlorine, Dicyan, Hydrogen chloride, Methanol,
Methyl bromide, Nitrogen trifluoride, Sulfur dioxide
The abbreviation LC50 stands for lethal concentration. Its value reflects the gas concentration in air which will kill 50 % of laboratory animals (mostly mice or rats)
when inhaled for a certain time (mostly four hours).
Toxic Gases and Vapours
Flammable Gases and Vapours
Flammable gases: The lower their lower explosion limit (LEL),
the more dangerous they are.
Flammable vapours: The lower their flashpoint, the more
dangerous they are. The flashpoint is defined by the liquid’s
temperature dependent vapour pressure and its LEL.
Vapour
LEL LEL
in
in
Vol-% g/m3
Acetone
Acrylonitrile
Benzene
n-Butanol
n-Butyl acetate
n-Butyl acrylate
Chlorobenzene
Cyclohexane
Cyclopentane
1.2-Dichloroethane (EDC)
Diethyl ether
1.4-Dioxane
Epichlorohydrin
Ethanol
Ethyl acetate
Ethyl benzene
n-Hexane
Methanol
1-Methoxy-2-propanol
Methylethyl ketone (MEK)
Methyl methacrylate
n-Nonane
n-Octane
n-Pentane
i-Propanol (IPA)
Propylene oxide
Styrene
Tetrahydrofurane (THF)
Toluene
Xylene (mix of isomers)
Gas
2.5
2.8
1.2
1.7
1.2
1.2
1.3
1.0
1.4
6.2
1.7
1.9
2.3
3.1
2.0
1.0
1.0
6.0
1.8
1.5
1.7
0.7
0.8
1.4
2.0
1.9
1.0
1.5
1.1
1.0
LEL
in
Vol-%
Acetylene
2.3
Ammonia
15.4
1.3-Butadiene
1.4
i-Butane
1.5
n-Butane
1.4
n-Butene (Butylene) 1.2
Dimethyl ether
2.7
Ethene (Ethylene)
2.4
Ethylene oxide
2.6
Hydrogen
4.0
Methane
4.4
Methyl chloride
7.6
Propane
1.7
Propene (Propylene) 1.8
60.5
61.9
39.1
52.5
58.1
64.1
61.0
35.1
40.9
255.7
52.5
69.7
88.6
59.5
73.4
44.3
35.9
80.0
67.6
45.1
70.9
37.4
38.1
42.1
50.1
46.0
43.4
45.1
42.2
44.3
LEL
in
g/m3
24.9
109.1
31.6
36.3
33.9
28.1
51.9
28.1
47.8
3.3
29.3
159.9
31.2
31.6
Flashpoint
in °C
< - 20
-5
- 11
35
27
37
28
- 18
- 51
13
-40
11
28
12
-4
23
- 22
9
32
- 10
10
31
12
- 40
12
- 37
32
- 20
6
25
Vapour
Ignition
pressure temperature
at 20 °C
in °C
in mbar
246
535
117
480
100
555
7
325
11
390
5
275
12
590
104
260
346
320
87
440
586
175
38
375
16
385
58
400
98
470
10
430
160
240
129
440
12
270
105
475
40
430
5
205
14
205
562
260
43
425
588
430
7
490
200
230
29
535
7
465
Ignition
temperatur
in °C
305
630
415
460
365
360
240
440
435
560
595
625
470
485
Only flammable
liquids have a
flashpoint.
By definition
there is no
flashpoint for
gases.
Flammable Gases and Vapours
LEL and Preventive Explosion Protection
Only if the proportion of a fuel in air is within certain limits
flammable gases and vapours can produce flammable mixtures
with air.
The lower explosion limit (LEL) is
defined as the concentration (given
in Vol%) of a fuel gas/air-mixture
which under standardized
conditions can be ignited and
continues to burn.
The LEL of all known flammable
gases and vapours is in the range
of approx. 0.5 through 15 Vol%.
For example, the LEL of hydrogenair-mixtures is 4 Vol%, and so a test
gas of 2 Vol% in air is definitely not
ignitable.
11.0 Vol%
Carbon monoxid
10.5 Vol%
10.0 Vol%
Formic acid
1.2-Dichloro ethylene
9.5 Vol%
9.0 Vol%
Methyl bromide
8.5 Vol%
8.0 Vol%
1.1.1-Trichloro ethane
Methyl chloride
7.5 Vol%
Concentration limitation
This behaviour is very essential for
the practice of explosion protection:
As a flammable gas below its LELconcentration cannot be ignited,
explosion protection can be
performed by continuously
measuring the gas concentration
and making sure by appropriate
measures that e.g. half of the LEL
(= 50% LEL) is never exceeded.
Acetyl chloride
7.0 Vol%
Formaldehyde
6.5 Vol%
1.1-Dichloro ethylene
6.0 Vol%
1.2-Dichloro ethane
Methanol
5.5 Vol%
1.1-Dichloro ethane
Hydrogen cyanide
5.0 Vol%
Methyl amine
Hydrazine
This method of preventive
explosion protection is known to
be the primary measure: Not the
ignition of a potentially explosive
atmosphere, but the forming of a
potentially explosive atmosphere is
reliably prevented.
Concentration measurement is
preferably performed by means
of infrared sensors or catalytic
bead sensors, which must comply
with special standardized safety
requirements and being certified
for this purpose.
4.5 Vol%
Methane
4.0 Vol%
Hydrogen cyanide
Vinylchloride
3.5 Vol%
Ethyl amine
3.0 Vol%
Ethanol
Aceto nitrile
Acrylo nitrile
Dimethyl ether
2.5 Vol%
Ethylene
Dimethyl formamide
2.0 Vol%
i-Propanol
Propane
1.5 Vol%
i-Butane
n-Butane
n-Butyl acetate
1.0 Vol%
0.5 Vol%
n-Hexane
n-Octane
n-Nonane
n-Decane
LEL and Preventive Explosion Protection
ST-3101-2004
Flashpoint of Flammable Liquids
Although speaking of flammable liquids, it is not the liquid
state but the vapour being flammable. Only vapour can form
a flammable mixture with the air’s oxygen.
Both the volatility of the vapour and its lower explosion limit
(LEL) are a measure for the risk of explosion. The properties
volatility and LEL are combined by the so-called flashpoint.
To be ignitable at all, the
concentration of the liquid’s vapour
above the liquid’s surface needs to
exceed the LEL. If it does or
doesn’t depends on the individual
vapour concentration produced by
the liquid’s vapour pressure – and
this is depending on the liquid’s
temperature only.
In respect to the safety of
flammable goods this behaviour
is described by the flashpoint (F):
The flashpoint is the temperature
at which just so much vapour is
produced that the vapour-airmixture can be ignited by a
standardized apparatus and
continues to burn. For example, if
the flashpoint of a flammable liquid
is above 50 °C, this liquid cannot
be ignited at 30 °C.
60 °C
Cyclohexanol
Dimethyl formamide
50 °C
Trimethyl benzene
Ethylglycol
40 °C
n-Butanol
30 °C
Nonane
Chlorobenzene
20 °C
Ethyl benzene
i-Butyl acetate
Ethanol
10 °C
Methanol
Toluene
0 °C
Aceto nitrile
Ethyl acetate
– 10 °C
Methylethyl ketone
Cyclohexane
– 20 °C
Hexane
– 30 °C
Allyl amine
You cannot ignite Diesel
(F > 55 °C) by means of
a burning match, but you
can ignite gasoline
(F < - 20 °C)!
Conclusion: flammable liquids are the more dangerous the
lower their flashpoint.
Since vapours of flammable liquids cannot be ignited below
their flashpoint, this may be the basis of preventive explosion
protection: Only use flammable liquids with flashpoints higher
than the ambient temperature and there is no risk of
explosion.
Indeed this is common practice, but if the liquids are used as
solvents there is a disadvantage: Less volatile liquids require
more energy for evaporation.
By definition, gases do not have a flashpoint because they
have no liquid phase under normal conditions.
Flashpoint of Flammable Liquids
ST-7448-2006
Gas Detection Systems
In a first approach gas detection instruments are products
of safety technology and are used preferably to protect
workers and to ensure plant safety. Gas detection systems
are dedicated to detect dangerous gas concentrations, to
trigger alarms and – as far as possible – to activate counter
measures, before it can come to a hazardous situation for
employees, assets and environment.
Quality assurance
Environment
System control
protection
Immission control,
Pollution control
Process control,
Profitability
ProProcess
zess m esste
chn ik
instrumentation
AnaAnalyzing
lyse nm ess
g erä te
instruments
Claim: Accuracy of measurement
Safety technology
Plant safety
Explosionsschutz,
Explosion
protection,
Hazard prevention
Störfallvermeidung,
Safety Integrity
Integrity
Safety
Safety at work
Personal protection,
Health and Safety,
Threshold Limit Values
sm esste chn
ik
Gas Ga
measuring
technology
swarn gerä
te
GasGa
detection
system
Claim: Reliability of warning
Gas detection instruments may be portable (or semi-portable)
gas measuring instruments or fixed installed gas detection
systems. The safety of an area potentially being affected by
dangerous gases and vapours to a high degree depends on
the reliability of the gas detection system, and especially on
the quality of the sensors being used. In opposite to sensors
of portable devices fixed installed sensors including their
electronics are continuously in operation year for year for
24 hours a day – just to be available for the instance of a
random gas release. And this even under extreme
environmental conditions, at e.g. - 50 °C or + 65 °C, at high
relative humidity or even very dry atmospheres, in outdoor
applications with rain, storm and snow or hot desert conditions,
electromagnetic disturbances or strong vibration ... And – selfevident – explosion protection must not be an issue and
measuring performance shall only be affected insignificantly.
As shown in the graphics, there is a fluent crossover between
gas detection technology on the one side and process
instrumentation on the other side. Although developed as
a product of safety technology, there are certain gas
detection transmitters having so excellent measuring
performance characteristics that today they even show up
more and more as analyzing instruments in the field of
process instrumentation.
Gas Detection Systems
ST-3751-2003
Measuring Principle Electrochemical Sensors
Many toxic gases are also very reactive and under suitable
conditions they change by chemical reactions. The
electrochemical sensor is such a micro-reactor, which at
the presence of reactive gases produces electrons just like
a battery. The flow of electrons is a very low but measurable
electric current.
ST-1101-2008
An electrochemical sensor consists of at least two electrodes
(measuring electrode and counter electrode) which have
electrical contact in two different ways: On the one hand via
an electrical conductive medium called electrolyte (a pastylike liquid to transport ions), on the other hand via an outer
electric current circuit (a simple copper wire to transport
electrons):
CO-sensor, schematic
Measuring
electrode
Counter
electrode
Micro-amp meter
CO molecule, target gas, penetrates into the measuring electrode
CO2 molecule, reaction product, departs from the measuring electrode
H2O molecule, part of the electrolyte
+
H , hydrogen ion with positive charge (because an electron is missing)
O, oxygen atom, aggressive, an electron-grabber
O2, oxygen molecule, decomposes to atoms on the counter electrode
electron, flowing electrons are measurable as electric current
The electrodes are made of a special material which also
has catalytic characteristics enabling certain chemical
reactions to take place in the so-called 3-phase zone,
where gas, solid catalyst and liquid electrolyte are present.
The electron grabber oxygen being needed for this reaction
comes from the ambient air. Further electron grabbers are
known, e.g. chlorine, fluorine, ozone or nitrogen dioxide.
Thus the sensor current of sensors being used for these
gases flows in reverse direction. The current can be
measured by means of a micro-amp meter.
Measuring Principle Electrochemical Sensors
Electrochemical Sensors
More than a hundred of gases and vapours are detectable by
electrochemical Dräger sensors. Some of these react very
specifically to the target gas, others are typical gas group
sensors being sensitive to a lot of different reactive gases.
Electrochemical Dräger sensors are mostly equipped
with three electrodes, a measuring-, a counter-, and a
reference-electrode. The sensor’s measuring performance
is increased by means of a bias voltage being measured and
kept constant by means of the reference-electrode and an
electronic control circuit (so-called potentiostat circuit).
Additionally there is a temperature measuring element in
the interior of the sensor because electrochemical
processes are extremely temperature dependent and
need to be compensated accordingly. Only by the outer
electrical circuit of the sensor (especially for temperature
compensation and amplification and conditioning of the very
low and noisy sensor current - only a few micro amperes)
which produces a 4-20-mA-signal, the electrochemical sensor
turns into a real gas detector. The following measuring
ranges can be realized (user adjustable with e.g. Polytron
7000):
Gas
Acidic Compounds
Ammonia
Carbon monoxide
Chlorine
Hydrazine
Hydrides
Hydrogen
Hydrogen chloride
Hydrogen cyanide
Hydrogen peroxide
Hydrogen sulfide
Nitrogen dioxide
Nitrogen monoxide
Organic Vapours
Ozone
Oxygen
Phosgene
Sulfur dioxide
minimum full
scale
deflection
3 ppm
50 / 300 ppm
50 / 200 ppm
1 ppm
0.3 ppm
0.3 ppm
500 ppm
20 ppm
10 ppm
1 / 1000 ppm
10 / 100 ppm
5 ppm
30 ppm
20 ppm
0.5 ppm
5 Vol%
0.1 ppm
5 ppm
maximum full
scale
deflection
30 ppm
200 / 1000 ppm
1000 / 5000 ppm
50 ppm
3 ppm
1 / 20 ppm
3000 ppm
100 ppm
50 ppm
50 / 7000 ppm
100 / 1000 ppm
100 ppm
200 / 500 ppm
100 / 200 ppm
5 ppm
25 / 100 Vol%
1 ppm
100 ppm
The electrochemical sensor needs very low electric power
and can thus be operated intrinsically safe. In this case
problems with heavy flameproof enclosures do not apply
and sensor replacement at site can be carried out very
easily and without hot work permit.
Electrochemical Sensors
ST-4940-2004
Measuring Principle Catalytic Bead Sensor
Under certain circumstances flammable gases and vapours
can be oxidized by means of the air’s oxygen to release heat
of reaction. Typically this is achieved by special and suitable
heated catalyst material, which slightly increases its
temperature by the heat of reaction. This increase of
temperature is a measure for the gas concentration.
So-called pellistors are tiny and very porous ceramic beads
(diameter approx. 1 mm) embedding a small platinum wire
coil. There is an electric current flowing through the platinum
wire coil so that the pellistor is heated up to some hundred
degrees Celsius.
If the ceramic bead contains some suitable catalytic material,
the pellistor’s temperature will rise in the presence of
flammable gas, and the platinum wire coil’s resistance will
increase accordingly. This change in resistance in respect to
the resistance in clean air is used for electronic evaluation.
ST-1581-2007
By means of the air’s
oxygen which is
adsorbed in the porous
material and activated
by the catalyst,
gaseous methane is
oxidized in the hot
pellistor. Besides water
vapour and carbon
dioxide measurable
heat of reaction is
released by this
reaction.
CH4 + 2 O2
2 H O + CO
2
2
+ heat of reaction
To eliminate influences of changes of the ambient
temperature, a second pellistor is used, which is very
similar but does not respond to gas, because the pellistor
does not contain catalyst material or is inhibited otherwise.
Integrating both the pellistors in a Wheatstone bridge
circuit results in a sensor for concentration measurement
of flammable gases, to a large extent independent of the
ambient temperature.
Measuring Principle Catalytic Bead Sensor
Catalytic Bead Sensors
One pellistor alone is not suitable for the detection of
flammable gases and vapours. It needs a second one to
compensate for environmental parameters (especially
temperature and humidity). And it needs to be explosion
protected. By means of a flameproof enclosure and a sinter
disk a useful catalytic bead sensor results.
ST-1103-2008
The compensator pellistor is built very similar to the active
pellistor, but does not contain catalyst material so that gas
cannot be oxidized. If the ambient temperature changes, the
resistance of both the pellistors will change and there is no
bridge signal. However, if gas is present, only the resistance
of the active pellistor changes and the Wheatstone bridge is
unbalanced. Since the pellistors of the catalytic bead sensor
are heated to about 450 °C, it can act as an ignition source
if the LEL is exceeded and the ignition temperature of the
gas is lower than 450 °C. By means of a sinter disk the
following is prevented: If in the interior of the catalytic bead
sensor an ignition takes place, the sensor’s housing will
withstand the explosion pressure and the flame is cooled
down below the ignition temperature of the gas, and no
The active pellistor and
the compensator are
placed in a flameproof
encapsulated housing.
The gas is penetrating
through the sinter disk
into the interior of the
sensor where it is
oxidized by the active
pellistor.
flame penetrates to the outside. These are the characteristics
of the so-called flameproof encapsulation.
Catalytic bead sensors are operated with an electronic circuit
called Wheatstone bridge, which is suitable to convert very
small resistance changes into a measurable voltage.
If the second half of the Wheatstone bridge is placed in the
controller, the catalytic bead sensor may be connected to
controllers by means of very long cables. However, they are
short when the sensor is installed in a transmitter.
Catalytic Bead Sensors
Measuring Principle Infrared Sensor
When considering the broad range of flammable gases and
vapours one will realize that most of these substances are
chemical compounds primarily consisting of carbon, hydrogen,
oxygen, and sometimes nitrogen. These so-called organic
compounds are called hydrocarbons. Hydrocarbons have
special properties which can be used for infraredmeasurement of their concentration.
ST-1104-2008
All the gases absorb radiation in a characteristic manner,
some even in the visible range (0.4 to 0.8 micrometers).
This is why chlorine is green-yellow, bromine and nitrogen
dioxide are brown-red, iodine is violet, and so on. However,
these colours can only be seen at rather high and lethal
concentrations. Hydrocarbons absorb radiation of a certain
wavelength range, approx. at 3.3 to 3.5 micrometers, and,
since oxygen, nitrogen and argon do not absorb, this can be
used for concentration measurement of hydrocarbons in air.
incoming
infrared intensity
reduced
infrared intensity
An excited methane molecule absorbs energy so it can vibrate
CH4 + energy
CH
4
(excited)
An optical system containing a mixture of e.g. methane or
propane in air will attenuate an incoming infrared intensity in
a predictable way, and for a given gas this attenuation is
depending on only its concentration.
Air: Infrared passes without being attenuated, no reduced
intensity, no measuring signal
Gas: Infrared passes by being attenuated, reduced
intensity, measuring signal corresponds to the current
gas concentration.
This photometer principle is the basis of an infrared
measuring instrument. The correlation of measured intensity
reduction on the one hand and the gas concentration in the
optical system on the other hand, is made by the calibration
process: A defined gas concentration will always produce
the same intensity reduction and thus always the same
measuring signal.
Most of the flammable gases and vapours are
hydrocarbons which are almost always detectable by
their characteristic infrared absorption.
Measuring Principle Infrared Sensor
Infrared Transmitters
The measuring principle is simple: Hydrocarbons absorb
infrared radiation (IR) in the wavelength range of 3.3 to 3.5
micrometers (μm), more or less, depending on the absorption
spectrum of the considered gas. However, the attenuation of
the infrared radiation is very small and a challenge concerning
the measuring technique. And, unfortunately, the reduction
of intensity can also occur by other circumstances, e.g. by
contaminated optics or reduction of the infrared radiation
source’s intensity.
The radiation source of an infrared sensor is a flashing
filament lamp, operated with low-voltage, having a high
percentage of infrared. Passing an IR-transparent window this
radiation is split into two parts by means of a beam splitter,
one part for the measuring detector, one part for the
reference detector. A detector consists of an encapsulated
pyro-electric crystal converting the received radiation energy
into a measurable voltage. However, the detectors are
different because of their optical interference filters: The
crystal of the measuring detector only receives radiation of
e.g. 3.4 μm, and the reference detector of e.g. 4.0 μm
wavelength. Gases generally do not absorb at 4.0 μm. So, if
both the detectors detect a reduced radiation energy, gases
surely are not the cause! By means of the reference detector
ST-1583-2007
it is possible to make the measurement to a certain degree
unsusceptible to contamination of the optics and even to
automatically request for preventive maintenance.
Encapsulation
IRsources
Beam
splitter
IRDetectors
Window
Reflector
Air- and gas molecules
IR-Sensor of the Polytron IR type 340, schematically
IR-Transmitters are equipped with IR sensors like this.
The measuring signal of both the detectors is conditioned
accordingly and is finally available as a 4 to 20-mA signal
at the transmitter’s output. Because of their long lifetime IRtransmitters are favoured in the industrial measurement.
In opposite to the electrochemical and catalytic bead
sensors the detectors of IR-sensors do not come into
contact with the gases to be measured. As long as there is
no condensation - and this is inhibited by heated surfaces –
IR transmitters perform so excellently that they can be found
more and more even in process industry applications.
Infrared Transmitters
ST-3719-2003
Diffusion Controlled Sensors
The high velocity of the gas molecules is the cause that
gases expand quickly and also quickly mix up with other
gases and never separate again.
And as long as there are concentration differences in the
whole mixture the process of mixing is incomplete and does
not come to an end.
ST-1105-2008
These concentration differences can also act as a micro
pump. If the concentration difference is kept constant there
will be a continuous flow of molecules into the direction of
the lower concentration – and this effect is used for sensors
in gas detection technology, the so-called diffusion-controlled
sensors.
The trick: With the catalytic bead sensor and the
electrochemical sensor the target gas is consumed by
chemical reaction, directly on the place of reaction we have
a gas concentration of nearly zero, far less then in the
ambient area, producing a depletion zone. So there is a
forced concentration difference and gas molecules are
flowing into the sensor’s reaction area.
Illustration: Because of the molecular movement the nature
does not keep calm until the three bigger molecules are
homogeneously dispersed over the entire gas volume. After
uniform distribution the diffusion process ends.
Convection to the sensor, diffusion into the sensor
While gas essentially gets to the sensor by means of natural
convection, the penetration into the sensor’s interior via
sinter disk or dust filter is rather a diffusion controlled
process, because pores contain calm air where no convection
takes place. Not the pore walls but the calm air is inhibiting
the penetration: If a gas molecule would be as big as a pea,
a sinter pore or filter pore would have a diameter of some
100 to 1000 meters!
Diffusion controlled sensors do not need a pump.
Diffusion Controlled Sensors
Open Path Detectors
If you think of an infrared-sensor with an optical path extended
for multiple times, this would be a gas detection system with
an open path: gas molecules entering this path will cause a
measurable IR-absorption. Gas detection systems like this are
something like a light barrier for gas molecules. And not only
for a few meters, this is true for open path lengths of up to
200 meters!
The measuring result (reading) of an open path
measurement is a little bit unusual. While the optical system
of an infrared-sensor can be assumed to be homogeneously
filled with a defined gas concentration, this is never the
case with a long measuring path.
Basically one cannot differentiate whether there is a high
gas concentration along a small part of the open path or half
the concentration along twice the path length: In both cases
the measuring signal is the same because each absorbing
molecule in the path contributes to the measuring signal,
independent from their distribution.
However, in special applications the probability that the gas
cloud is drifting through the open path is higher than being
detected by a point detector.
ST-1108-2008
In the terms of safety: One abandons the usual concentration
measurement for the sake of higher detection probability.
Illustration: As long as the gas cloud stays in the open path
there is a measuring signal, independent of the dilution with
air. We may say an open path detector is a molecule counter
independent of their distances.
So, an open path detector is rather a reliable gas hazard
indicator than a concentration measuring instrument. In
applications handling a high risk of gas release (e.g. natural
gas exploration) the information „gas is present“ is considered
to be sufficient to activate safety relevant counter measures.
Open Path Detectors
4 to 20 mA Transmitters
A stand-alone sensor is not sufficient for gas detection.
Sensor signals need to be electronically conditioned (some
of them temperature-compensated), locally displayed (for
calibration purposes), measuring ranges to be configured –
and for all this power is necessary.
It has become worldwide accepted that gas detection
transmitters are operated by 24 V DC, and this voltage may
vary within wide limits, e.g. between 16 and 30 volts. The
transmitter’s electronics convert the sensor signal into an
output current so with clean air (zero point) at the sensor a
current of 4 mA flows to the control unit, while it is 20 mA
at full scale deflection.
If the cable between transmitter and controller is cut, no
current will flow (the current is 0 mA), and the controller
will recognize this condition, activating a fault alarm.
Moreover, signals lower than 3.8 mA or higher than 20.5 mA
are not interpreted as a measuring signal but as special
signals to indicate underrange (negative measuring values)
or overrange, and also special maintenance signals. The 4 to
20 mA-signal is a worldwide accepted industry standard
which, in opposite to voltage signals, does not depend on the
cable resistance, which has low impedance (resistance) and
is relatively immune to electromagnetic interferences.
Transmitters with electrochemical sensors have such a
low power consumption that they can be operated with
even less than 4 mA, and can draw, depending on the gas
concentration, an additional current of 4 to 20 mA from the
power supply: Only two wires are needed for power supply
and measuring signal. However, transmitters with catalytic
bead sensors or infrared-sensors have higher power
consumption and require 3 wires for power supply and signal.
The 4 to 20 mA signal can also be used to superpose a
symmetrical frequency modulated signal of ± 0.5 mA to
exchange additional digital information between controller
and transmitter.
This is the so-called HART signal, which can also be used to
address certain transmitters by their digital addresses and
poll their actual measuring signals and more. Several HARTcompatible transmitters can communicate on only one 2-core
communication cable.
Digital communication, via HART or via the so-called
RS 485 interface, enables the customer to remotely query
the transmitters’ “well-being” and preventively perform
maintenance actions based on the results.
4 to 20 mA Transmitters
Sampling
There are many reasons not to have the sensor directly at
the location where the target gas (gas to be measured) may
occur. Instead of this continuous sampling can be performed
which may also have advantages: The gas sample can be
pre-conditioned (e.g. by filters), can be dried and tempered,
condensate can be trapped, pressure changes can be
compensated, and by means of solenoid valves further
sampling paths can be activated. Especially test gas can
be applied automatically to the sensor.
Zone spreading
When sampling from a hazardous area (mostly zone 1) the
ex-zone is spread via the sampling tube into the safe area.
Since in the safe area there are no means of explosion
protection, ignition becomes possible. This can be avoided by
using flame arrestors in the sampling tube. Flame arrestors
do not inhibit the ignition, but prevent a flash-back into the
hazardous area.
Sampling tubes
The smaller the sampling tube’s cross section, the higher the
pressure drop – the greater the tube’s cross-section, the
longer the response time: 4 mm inner diameter and a flow of
about 1 to 2 liter per minute is a good compromise.
Pre-sampling
Using a strong pump (10 to 20 L/min) gas sampling can be
realized over longer distances (up to 100 to 150 m). A second
small pump (approx. 1 L/min) should be used to supply the
gas sample taken from the pre-sampling tube to the sensor.
Sampling line monitoring
Considering the safety relevance it is essential to know that
sampling basically increases the response time and that the
complete sampling line needs to be monitored for proper
function. This is mostly done by a flow meter with switch
contacts which, in case of a blocked sampling tube or pump
failure, trigger an alarm. If filters or condensate traps are
used, they need to be maintained regularly.
Material and wall adsorption effects
The tubing material should preferably be PTFE (Teflon),
Viton, or stainless steel. Some gases tend to increased
adsorption on the material’s surfaces and this may influence
the measuring signal especially for ppm-measurement.
Sampling
Explosion Protection
In industrial processes very frequently flammable substances
and sometimes also flammable dusts are involved. In these
areas flammable gases and vapours may be released
process-related (e.g. by relief valves) but also by
unpredictable and hazardous incidents. Preventively these
hazardous areas are declared to Ex-areas (“zones”), in which
solely equipment may be installed which is equipped with a
suitable type of explosion protection and certified accordingly.
Explosion protection is worldwide regulated. The basis of
these standards according to IEC, CENELEC (Europe) and
NEC 505 (North America) is very similar and is established
on the “3-zones-concept“, which is also more and more
accepted in the USA,
Zone according to
IEC, NEC 505
and CENELEC
Zone 0
Zone 1
Zone 2
Dangerous explosive
atmospheres are present ...
continuously, long periods
or frequently
occasionally, likely to occur
infrequently and for a short
period only
while the well-known American way of explosion protection
acc. to NEC 500 is based on the “2-divisions-concept“:
Division acc. to
NEC 500
Division 1
Division 2
Dangerous explosive
atmospheres are ...
likely to exist
not likely to exist
According to IEC, NEC 505 and CENELEC there are seven
standardized types of protection for electrical equipment in
zone 1, while in North America (USA/Canada) there are only
three types of explosion protection for division 1 according to
NEC 500:
Type of protection
acc. to IEC, NEC 505
and CENELEC
Flameproof enclosure
Encapsulation
Powder filling
Oil immersion
Pressurized apparatus
Increased safety
Intrinsic safety
comparable type of
protection acc. to
NEC 500
Explosion proof
–
–
–
Purged / Pressurized
–
Intrinsically safe
Nowadays powder filling and oil immersion are scarcely
being used in the measurement and control technology and
are not at all applied in gas detection. Pressurized apparatus,
this means continuous purging e.g. with pressurized air, is a
typical type of protection for major devices and cabinets.
The standardized marking of an explosion protected device,
e.g. Ex de IIC T4 or Class I, Div1, Group B, C, D, informs
the expert about the applicability in the designated
hazardous area.
Explosion Protection
STL-296-2007
Type of Protection Intrinsic Safety
Products for measuring and control technology, which have
relatively low power consumption, can be designed very smart
in respect to their explosion protection. Smart, because the
product is designed such that even in case of a first or
second failure it is ensured that neither sparks of sufficient
energy nor surfaces of sufficient temperature can occur – so
no risk of ignition.
Electrical discharges must have a certain minimum energy
(ignition energy), otherwise they will not be able to ignite
the most flammable mixture of a given flammable gas with
air – hence, certainly no other arbitrary mixture of this gas in
air can be ignited.
Also, for a given gas, the surface temperatures of electric
or electronic components shall not exceed a certain
temperature (ignition temperature).
Therefore, if the electronic circuits of a product are designed
accordingly and the stored electric energy (i.e. effective
capacities and inductivities) as well as electrical power (i.e.
electrical current and voltage) are limited to certain maximum
values, this electronic circuit cannot act as an ignition
source – the product is said to be intrinsically safe.
There is an important accessory for this: When intrinsically
safe current circuits lead into the hazardous area they need
to be protected against too high power by so-called safety
barriers. Safety barriers at least contain a fuse, resistors for
current limiting and Zener diodes for voltage limiting.
Intrinsically safe products are marked by an “i”.
Their design is sophisticated, lightweight and simple – and
intrinsically safe products can be maintained while energized.
Cables may be disconnected and sensors may be replaced
without zone declassification – because sparks and hot
surfaces surely cannot exist.
Type of Protection Intrinsic Safety
ST-9135-2005
Type of Protection Flameproof Enclosure
The type of protection “flameproof enclosure“ is the oldest of
all, having been used in the early mining industry since the
beginning of last century (the marking “d“ refers to the
German origin of this way of explosion protection). Compared
to the protection type intrinsic safety a flameproof enclosure is
a heavy and purely a mechanical method to avoid the ignition
of an explosive atmosphere:
Simply let the explosion take place in the interior of the
enclosure and reliably avoid a flashback. So housings of the
protection type flameproof housing must be designed such
that they will withstand the inner explosion pressure. The
greater the housing’s volume, the higher is the possible
explosion pressure, the more robust the housing shall be.
If flammable gases penetrate into the interior of the housing,
one can assume that the housed electronic circuits (which
also may produce sparks or have hot surfaces) will ignite the
gases. On ignition the enclosure will withstand the explosion
pressure, and the explosion pressure is released via joints.
These mostly metallic joints with a certain minimum surface
(with defined gap width and gap length) have an important
function: Hot gases flowing along these joint paths are
cooled down below their ignition temperature – a very
effective flame extinguishing. With this type of standardized
protection a potential flame in the enclosure’s interior cannot
flash back into the hazardous area.
Flameproof housings are robust and heavy, and they are
not allowed to be opened when energized. For maintenance
an official hot work permission is necessary. The electrical
connection of flameproof enclosed instruments can be made
in three different ways:
1. Rigid Conduit: The electrical wires are run inside
enclosed approved metal pipes. The pipes are directly
screwed into the tapered NPT thread of the instrument.
The complete conduit system is specially sealed and
flameproof.
2. The cable is connected via an approved flameproof
cable gland. Disadvantage: Explosion protection cannot
be ensured by the manufacturer but only by the installer
on-site.
3. The cable connection is done via an approved junction
box with increased safety (“e”).
Type of Protection Flameproof Enclosure
ATEX 95 – European Directive 94/9/EG
Also known as ATEX 95 (formerly ATEX 100a), mandatory in
the European countries (EU) since 1st of July 2003, addressed
to the manufacturers. Equipment and protective systems for
the use in potentially explosive atmospheres need to fulfill the
Essential Health and Safety Requirements (EHSR) which are
assumed to be met when based on certain harmonized
standards.
CE-marking for free trade within the European Union:
0158
Notified Body Number concerning quality
EU-requirements are met
Marking (according to ATEX):
II 2 GD
Type of potentially explosive atmosphere:
G: Gas, vapour; D: dust
Category
I: Mining, II: other areas except mining
complies with the directive 94/9/EC
Explosion protection:
Ex ib IIC T4
Temperature class
Explosion group: I: Mining, II: other areas except
mining for ia, ib, d and n: Subgroups IIA, IIB und IIC
Type of protection (here: intrinsically safe)
Explosion protected equipment
EC-Type Examination Certificate:
TPS 04 ATEX 1003X
X: Special conditions
U: incomplete Ex-component
Number of certificate
complies with the directive 94/9/EC
Year of the certificate’s publication
Notified Body having type-approved the equipment
Device categories and safety requirements:
Device group
I (Mining)
II (other areas except mining)
Category
M1
M2
1
2
3
Safety
very high
high
very high
high
normal
ATEX 95 – European Directive 94/9/EG
ST-334-2008
ATEX 137 - European Directive 1999/92/EG
Also known as ATEX 137 (formerly ATEX 118a), mandatory
in the European countries (EU) since 30th of June 2006,
addressed to employers and end-users concerning the
minimum requirements for health and safety for workers in
potentially explosive atmospheres.
Zone definition:
Gas, vapour Dust
zone 0
zone 20
zone 1
zone 2
zone 21
zone 22
Explosive atmosphere
continuously, long periods or
frequently
occasionally, likely to occur
infrequently and for a short
period only
Selection of equipment (this table is the link between
categories of the ATEX 95 and zones of the ATEX 137):
Operation allowed
Devices of category
Devices of category
Devices of category
for
1
2
3
Gas, vapour (G) Dust (D)
in zone 0, 1, 2
in zone 20, 21, 22
in zone 1, 2
in zone 21, 22
in zone 2
in zone 22
Example: In zone 21, where explosive atmospheres caused by
dust are likely to occur, the instruments to be used need to
have a marking II 2D or II 1D.
Necessary measures:
- Assessment of the risk of explosion
- Classification of the hazardous area into zones
- Marking of the hazardous places by means of a triangular
warning sign “Ex”
- Adequate safety measures
- Explosion protection document
- Competence of employees
- Criteria for a permit-to-work system for dangerous work
Guideline for risk reduction:
- Prevent the formation of explosive atmospheres,
or, if this is not possible:
- Avoid the ignition of the explosive atmosphere,
or, if this is not possible:
- Minimize harmful effects of explosions to a tolerable
degree.
ATEX 137 – European Directive 1999/92/EG
Safety Integrity – SIL
The term Safety Integrity seems to be something like a fashion
in the technical-based and automatically operated safety
technology. But it is not, and fixed installed gas detection
systems are not exempted from this trend to classify protection
systems by means of the so-called Safety Integrity Level (SIL) in
respect to their reliability to activate a risk-reducing safety action.
- Wherever there is a technical risk for people, assets or
environment it is necessary to gain a certain degree of
safety by means of risk reducing measures. If such
measures are realized automatically by electric, electronic
or programmable electronic systems the central term is
“functional safety“.
- Such systems, frequently called protective systems or
safety relevant systems perform a safety function and need
to be adequately reliable in respect to the actual risk.
- However, such systems can fail by any arbitrary failure. If
they fail they must not remain in a dangerous (unsafe)
state and need to be repaired instantly. This however
implies that the occurrence of a failure is detectable at all.
- A high fraction of all the possible failures (which are
identified by a so-called FMEDA) can be made detectable
by failure monitoring (diagnostic facilities), so that the
-
-
-
-
system in case of a detectable failure can be forced into
the safe state, which is said to react fail-safe.
Statistically there is a very small remaining fraction of
accidental dangerous failures which cannot be detected
automatically, the so-called dangerous undetectable or
DU-failure, which might be seldom, but will impede the
execution of the safety function.
Setting the probability of the occurrence of a DU-failure
into relation to all possible failures this will result in the socalled Diagnostic Coverage value DC and the important
Safe Failure Fraction SFF. Both these values must
exceed certain percentages depending on the individual
safety requirements.
By ingenious system concepts (especially by
redundancies), periodically repeated function proof tests
and preventive measures the probability of the occurrence
of a DU-failure can additionally be decreased.
The remaining residual risk can be assessed statistically
and classified. This results in four different safety integrity
levels SIL1 to SIL4, where SIL4 is reflecting the highest
reliability but is not established for gas detection systems,
which are rather SIL2 and – by means of redundancies –
SIL3 rated.
Safety Integrity – SIL
ST-899-2007
Alarm Philosophy
What to do in case of an alarm? Gas detection systems are
designated to trigger alarms in sufficient time to inform the
operator about the alarm condition and to perform counter
measures for the prevention of a dangerous situation. This
is mostly an automatism, but can also be achieved by
establishing an alarm plan. It is in the operator’s responsibility
to adequately react in case of an alarm.
The safety concept of a gas detection system always is:
Detect dangerous gas, react and avert.
Main alarm
The exceeding of only one alarm threshold (main alarm
threshold) is basically sufficient. By this the safe state is
achieved by protecting the hazardous area (visible / audible
alarm and evacuation) or by shutting off gas supply or by
disable the ignition sources, call for use of personal protection equipment or breathing protection etc. This is a safe but
rigorous, so to say uneconomic measure – the entire process
is affected and shut-down.
Pre-alarm
This can possibly be avoided by means of a pre-alarm which
is activated at lower concentrations than the main alarm
threshold. With the pre-alarm automatic counter measures
can be initiated which, if they are effective, will prevent the
main alarm from being triggered, e.g. a pre-alarm can activate
an effective ventilation so that the gas concentration stops
rising and the main alarm threshold is not exceeded. This is
ideal because: Via a pre-alarm a dangerous situation can
be controlled without process shut-down. It is the
operator’s interest to design counter measures so effective
that the main alarm will most likely never be triggered:
Properly designed gas detection systems will reach
main alarm only seldom or never.
Fault condition alarm
Fault condition alarms indicate that the system is partly or
entirely inoperative and in case of a gas release cannot react
properly. Preventively the same measures need to be taken
in case of fault condition as in case of a main alarm, since
there is no gas detection system at all. With this philosophy
a safe condition is achieved best.
Alarm Philosophy
ST-725-2002
Sensor Positioning
A gas detection system’s reliability is not only depending on the
properties and performance of the equipment, but also depending on installation, operation and maintenance – and especially
the proper positioning of sensors. Naturally, sensors can only
detect a gas when the sensor is within the gas cloud. Incorrect
sensor positioning results in a useless gas detection system.
Gas leaks can arise e.g. when cold liquefied and/or
pressurized gases are released into the ambient air, and
get mixed with it. Their concentration decreases, and the
gas dispersion depends more on the current temperature
conditions and ambient air convection than on the density
of the pure gas.
Three Rules of Thumb:
- There are only three flammable gases which are
considerably lighter than air: Hydrogen (H2), Ammonia
(NH3), and Methane (CH4). Commonly mixtures of these
gases rise up.
- Vapours of flammable liquids are always heavier than air –
they flow downwards as long as they are not disturbed by
air convection.
- Independent of the density of the pure gas, gas
concentrations of less than 1000 ppm in air virtually have
the same density than air. Dispersion of concentrations like
this will rather follow the current temperature profile and air
convection.
Positioning strategy:
Having the sensors as close as possible placed at the
potential leak surely is the optimum way. Leaks may arise
from pumps, valves, flexible tubes and their connections,
flanges, shut-off devices, bellows, etc. If such locations
cannot easily be identified, sensors need to be distributed
over the entire hazardous area (area monitoring). It is
essential that the target gas always can reach the sensor at
operational conditions within a given time interval. The local
conditions of the individual hazardous areas are so very
different that there are no regulatory standards where to
place a sensor, but useful guidelines exist (e.g. the EN
50073 or IEC 60079-29-2).
Sensor Positioning
ST-1005-2004
Calibration
In a first approach gas sensors do not measure gas
concentrations directly: Electrochemical sensors measure
electron flow changes, catalytic bead sensors measure
resistance changes, and IR-detectors measure changes of
IR-radiation intensity in the near infrared. These changes
always refer to the normal condition (clean air) which is
called zero point because no target gas is present. Only
by calibration it is possible to correlate a certain gas
concentration to a certain output signal, resulting in a
gas detection instrument.
Calibration is extremely important. Obviously, gas detection
instruments cannot measure properly as long as they have
not been calibrated properly. While zero-calibration is rather
simple because ambient air can mostly be used for this
purpose, calibration of the sensitivity (called span-calibration)
is not so trivial.
For the same reason that electrochemical sensors can
detect reactive gases, they have to be calibrated using
reactive gases. But unfortunately a lot of reactive gases also
react with (moist) material surfaces and plastics.
Although from the aspect of safety it is recommended to
perform the span-calibration with the target gas (the gas
which shall be detected) there are several reasons to use an
easy-to-handle surrogate test gas for cross-calibration.
If a variety of gases or vapours shall be detected by only one
sensor, the instrument has to be calibrated for the substance
the sensor is the least sensitive to. Thereby, the gas detector
is calibrated to the safe side, because all gas concentrations
are either measured properly or as too sensitive. The
sensor’s sensitivity for certain gases cannot be calculated
from gas specific data, but can only be determined by
applying the gas and evaluate the response.
To achieve good measuring performance, calibration should
be performed as good as possible under the expected
conditions during operation.
Calibration chamber for flammable liquids
To obtain a given %LEL concentration of flammable vapours
it is recommended to use a calibration chamber, where a
certain calculable amount of liquid (e.g. 100 microlitres) has
to be inserted. After complete evaporation a concentration of
e.g. 50 %LEL is formed which can be directly applied to the
sensor.
Calibration
Requirements for Gas Detection Systems
Since gas detection instruments and systems are products of
safety technology for industrial applications they need to
comply not only with statutory requirements (e.g. electrical
safety, explosion protection, electromagnetic compatibility)
but also with further requirements such that even in harsh
industrial environments the product’s quality and reliability
of alarming will sustain.
Standards concerning explosion protection:
Design requirements make sure, that devices will not act as
a source of ignition. World-wide accepted standards are
issued by e.g. CENELEC (ATEX), IEC, CSA, UL, GOST, etc.
Electromagnetic compatibility acc. to EN 50 270:
Test standards ensure that devices do not produce wire
bound or radiated disturbances, and especially, that they
are not negatively influenced by wire bound disturbances
(surges, bursts) or high frequency emissions (80 MHz to
2 GHz at field strengths up to 30 V/m), and that they
withstand electrical discharges without being affected in
respect to their reliability. The relevant test standards are
based on the series IEC 61000.
Climate, vibration and impact, e.g. acc. to IEC 60028:
Dry and damp heat temperature cycles (up to 70 °C,
including condensation!) over several days and cold test.
During these cycles, sporadic function tests and insulation
resistance tests with high voltage are performed. Vibration
test up to the 4-fold gravity for 90 minutes each axis at the
determined resonance frequencies have to be passed without
affecting proper function.
Measuring performance:
A certain measuring performance has to be met even under
extreme environmental conditions (temperature, pressure,
wind speed, relative humidity, vibration, etc):
EN 61 779 / EN 60079-29-1 – for flammable gases/vapours
EN 45 544 – for toxic gases and vapours
Digital communication acc. to EN 50 271:
Transmitter and controller nowadays are mostly microprocessor-controlled. Hardware and software need to comply with
certain requirements concerning the system’s reliability.
Shipping approvals:
Requirements of the so-called classification societies e.g. Det
Norske Veritas (DNV), Lloyds Register of Shipping (LRS),
Germanischer Lloyd (GL), Bureau Veritas (BV), etc.
Requirements for Gas Detection Systems
STL-1106-2008
Properties of dangerous
Gases and Vapours
Flammable and toxic gases and vapours may occur in
many places. To deal with the toxic risk and the danger
of explosion – this is what gas detection systems are for.
This brochure is meant to give a basic introduction to
gas detection technology, measuring principles and the
safety concerns.
Sensors and
Measuring Principles
Sensors for the detection of gases and vapours are
transducers making use of certain properties of gases
for the conversion into a suitable electrical signal.
Especially three measuring principles have become
mainstream in the recent decades of industrial gas
detection: Electrochemical sensors, catalytic bead
sensors and infrared sensors.
Explosion Protection
and Plant Safety
In almost every industrial application flammable
substances are used. Equipment installed in these
areas needs to be explosion protected and must be
approved and certified for this purpose.
With suitable safety concepts, high availability and
failure resistance, gas detection systems turn into
reliable protection systems.
Dräger Safety AG & Co. KGaA
Revalstrasse 1
23560 Luebeck, Germany
Tel +49 451 882 0
Fax +49 451 882 2080
www.draeger.com
FRANCE
UNITED KINGDOM
Dräger Safety France S.A.S.
3c, Route de la Fédération
67025 Strasbourg Cedex
Tel +33 388 40 76 76
Fax +33 388 40 76 67
Draeger Safety UK Ltd.
Blyth Riverside Business Park
Blyth, Northumberland NE24 4RG
Tel +44 1670 352 891
Fax +44 1670 544 475
P. R. CHINA
USA
Beijing Fortune Draeger
Safety Equipment Co., Ltd.
Yu An Lu A 22, B Area
Beijing Tianzhu Airport
Industrial Zone
Houshayu Shunyi District
Beijing 101300
Tel +86 10 80 49 80 00
Fax +86 10 80 49 80 05
Draeger Safety, Inc.
505 Julie Rivers
Suite 150
Sugar Land, TX 77478
Tel +1 281 498 1082
Fax +1 281 498 5190
SINGAPORE
Draeger Safety Asia Pte. Ltd.
67, Ayer Rajah Crescent # 06 03
139950 Singapore
Tel +65 68 72 92 88
Fax +65 67 73 20 33
90 46 421 | 08.08-1 | Marketing Communications | CR I PR | © 2008 Dräger Safety AG & Co. KGaA
SYSTEM CENTERS