Mettler Toledo Hints and Tips to Improve Accuracy In DO and pH Measurement (Reference / Paper)
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Accuracy Guide
Operating Principles and
Practice for Optimal
Performance
Optical Dissolved
Oxygen Sensors
pH Sensors
Quality Measurements for
In-line Pharma Applications
Hints and Tips to Improve Accuracy
In DO and pH Measurement
Contents
1. Introduction
1
2. Accuracy and precision
1
3. Calibration
3
4. H
ow accurate is a pH measurement?
6
5. Accuracy of dissolved oxygen sensors
9
6. Conclusions
12
1. Introduction
High quality and efficacy of products and efficiency in production are the main goals in pharmaceutical manufacturing. To achieve these aims, production processes must be stable, predictable and operate consistently at
the target level of performance.
Process analytics using in-line sensors plays a major role in monitoring manufacturing to ensure the required
process conditions are always being met. Obviously, reliability of the data from sensors correlates with sensor
measurement accuracy.
Accuracy is not solely dependent on the use of high quality sensors. How sensor maintenance and calibration
are conducted significantly influence a probe’s ability to measure dependably. In fermentation or cell culturing
processes, assurance that sensors will output reliable measurements throughout a batch is particularly critical.
This guide covers good operating procedures and includes advice for maintaining in-line pH and dissolved
oxygen (DO) sensors to ensure high measurement reliability at all times.
2. Accuracy and precision
Whereas accuracy is the proximity of measurement results to the true value; precision is the reproducibility of
the measurement.
probability
density
accuracy
trueness
value
reference
value
measured
value
If a measurement is precise it does not mean that it is also accurate. Depending on the required precision,
the opposite is also true. High accuracy can therefore be defined as a true value that is reported with great
precision.
Hints and Tips to Improve Accuracy
METTLER TOLEDO Process Analytics
1
In fermentation or cell cultures often only one pH or DO sensor is installed. It is not possible to determine measurement trueness in such situations (unless an off-line measurement is taken). However, a pH reading to one
decimal place is sufficiently precise for most applications. Only if two or more pH or DO sensors are installed
redundantly can we have confidence in the transmitted values. Double measurement is therefore more common
in demanding mammalian cell cultures.
Normally in biopharma processes, reference pH or DO values are not highly precise. There is an acceptable
range around these values that is determined by regulations (e.g. the USP) or the manufacturer. The concept of
Quality by Design (QbD) is based on acceptable ranges rather than a single reference value. This design space
philosophy enables flexibility in manufacturing.
An acceptable range suggests that sensor accuracy is not critical. However, if a sensor is reporting erroneous
values, it is not possible to determine whether a measurement is within or not within the acceptable range.
Performance
parameter
(product quality etc.)
Process validation
acceptance range
Design space
Operating space
Operational parameter
(pH, DO, etc.)
Acceptable range
2
Hints and Tips to Improve Accuracy
METTLER TOLEDO Process Analytics
Acceptable range according to QbD
Example: pH control during fermentation
In cell cultures and fermentation processes, optimal growth conditions are necessary throughout a batch to
prevent low yield, a slower batch time, or the production of unwanted byproducts.
Cell conc.
Lactate conc.
pH 7.0
pH 7.3
pH 7.3
time
pH 7.0
time
As can be seen in the graphs, maintaining pH 7.0 increases cell concentration and reduces lactate concentration (lactic
acid influences the quality of final product and make purification steps more difficult) compared with pH 7.3.
3. Calibration
Calibration is the comparison of a calibration standard of known accuracy (e.g. pH buffer) with another instrument of unknown accuracy (e.g. a pH sensor) to detect, correlate, report or correct any variation in the accuracy
of the item being compared (e.g. a pH transmitter).
Hints and Tips to Improve Accuracy
METTLER TOLEDO Process Analytics
3
Benefits of calibration
pH and DO sensors have a certain accuracy associated with them as provided by the manufacturer. This accuracy can only be maintained if sensors are regularly and correctly calibrated. Further, calibration is essential in
ensuring compliance with regulations such as global pharmacopeia.
Issues with sensor calibration
A common procedure with pH sensors is as follows: A calibrated pH sensor is mounted on a bioreactor. After bioreactor sterilization, filling with nutrient media and inoculation, the batch can be started. For verification of the measured pH
value a sample is taken and sent to the laboratory for measurement. If the pH measured in the laboratory differs from
the in-line pH measurement in the bioreactor, operators adjust the pH transmitter to perform a process calibration.
The above procedure is formulated in approved Standard Operating Procedures (SOPs). However, it is only valid
if the lab pH measurement is correct. To achieve this, the sample in the laboratory must be measured at the
same temperature that is in the bioreactor, and the time between sample and lab measurement must be minimal. For many pharmaceutical companies, this is counter to their actual process. In some situations the lab
measurement may be conducted at ambient temperature or thermo-controlled at 25 °C. In other situations the
laboratory will make use of the fact that most pH sensors contain a temperature sensor and that the connected
meter or transmitter will allow temperature compensation.
However, for a proper calibration and therefore an accurate pH measurement, sample temperature should be
equal to process temperature. The reason for this is that there are two independent temperature dependences
in pH the measurement: temperature dependence of the chemical equilibria in the medium, and temperature
dependence of the pH sensor.
Chemical equilibria are temperature dependent
Chemical equilibria are dynamic and therefore respond to changes in the conditions. Equilibria can be expressed with the equilibrium constant K, which is temperature dependent.
T (°C)
pH + 0.02
T (°C)
pH + 0.02
0
7.12
50
6.97
5
7.09
55
6.98
10
7.06
60
6.98
15
7.04
65
6.99
20
7.02
70
7.00
25
7.00
75
7.02
30
6.99
80
7.04
35
6.98
85
7.06
40
6.97
90
7.09
45
6.97
95
7.12
Buffer Solution pH 7.00
Temperature dependence
The temperature dependence of the media in the bioreactor is normally not known. What is well known however is
the temperature dependence of buffer solutions. Each label on pH buffer bottles shows the exact value at different
temperatures.
4
Hints and Tips to Improve Accuracy
METTLER TOLEDO Process Analytics
Most transmitters contain stored pH calibration buffer data and the chosen buffer must be selected on the
transmitter for a proper calibration. Automatic temperature compensation in transmitters and lab meters has no
connection with the temperature dependence of chemical equilibria.
The output of a pH sensor is temperature dependent
The internal operations of a pH sensor is based on potentiometric principles. The output of a combination pH
sensor (pH and reference electrodes) is primarily the potential difference between the pH-sensitive glass and
the reference electrode. This potential difference, also called cell potential, is measured in volts. The temperature
dependence of this cell potential is described by the Nernst equation.
E = E0 + 2.303 x R x T / F x log (aH+)
or
E = E0 - 2.303 x R X T / F x pH
the factor 2.303 x R x T / F is called the slope (V/pH)
R = gas constant 8.314 J K-1 mol-1
F = Faraday constant 96485 A s mol-1
Examples:
At 25°C (298.15 °K) slope = 2.303 x 8.314 x 298.15 / 96485 = 0.05917 V/pH or 59.17 mV/pH
At 35 °C slope = 61.15 mV/pH
The calculated slope according to the Nernst equation can also be expressed as a 100% slope. An actual pH
sensor shows a smaller slope. End control criteria for a new pH sensor from METTLER TOLEDO is >98% slope.
After exposure in a process, slope will decrease slightly.
The temperature dependence of the slope at different temperatures is compensated for automatically in the pH
transmitter or lab pH meter. This automatic temperature compensation (ATC) occurs during calibration and
measurement and is driven by the built-in temperature sensor. The ATC function can also be deactivated.
The above means that temperature compensation of the slope has no correlation with the temperature dependence of the chemical equilibria of a given media.
Summary
For correct pH comparison with in-line versus off-line equipment, the measuring temperature must be identical.
In some cases the temperature coefficient is quite small and correct measurement can be achieved even at
different measuring temperatures.
Practical hints
Measure a sample in the lab at different temperatures with activated ATC. If you do not see any difference in the
displayed pH, you can safely compare the measurement at different temperatures. This is only valid if media
composition is unchanged from batch-to-batch.
Hints and Tips to Improve Accuracy
METTLER TOLEDO Process Analytics
5
Samples from a bioprocess often contain CO2. If CO2 escapes from the sample a pH shift can occur. Therefore,
good sampling technique and quick lab pH measurement are essential.
4. How accurate is a pH measurement?
Structure of a “classical” pH measurement loop
A classical pH loop consists of a combination pH sensor, connected with a cable to a pH instrument (pH meter,
pH transmitter).
A classical pH sensor comprises a pH sensitive glass electrode and a reference electrode. The gel layer build-up
on the pH sensitive glass generates a pH dependent potential which can be measured behind the pH-sensitive
glass (sometimes called the glass membrane) in a defined inner buffer. A lead-off element transports this
potential out of the electrode.
HA465-50-S7 analog pH sensor
The reference electrode, containing reference electrolyte, is connected through a liquid junction (e.g. ceramic
diaphragm) to the media or sample. Ideally, the potential of the reference electrode is pH independent.
As mentioned, the potential difference between pH glass electrode and reference electrode is the primary signal
output by a combination pH sensor. This signal is of very high impedance (100-500 MOhm) and with a voltage
< 1/- 800 mV. Most pH sensors are designed to generate a voltage of 0 mV at 25°C at pH 7. The pH dependence of this voltage can be described by the Nernst equation. Further, this equation explains the temperature
dependence of pH sensors. This temperature dependence can be compensated for if the pH instrument is
equipped with the aforementioned automatic temperature compensation (ATC), ideally based on a temperature
sensor built-in the combination pH sensor. It is a common misunderstanding that ATCs do not correct for the
change in solution pH with temperature.
6
Hints and Tips to Improve Accuracy
METTLER TOLEDO Process Analytics
The pH instrument must have an input resistance greater than 1 TeraOhm. With today’s electronics this is a
much easier task than it was 50 years ago. The pH instrument converts the mV signal to a pH value but is also
used for calibration and adjustment based on measurement in pH buffer solutions. The expression “pH sensor
calibration” is misleading. It is not the sensor which is calibrated, it is the attached instrument (transmitter or
meter) which is adjusted to the output of the pH sensor.
Accuracy of a standard pH sensor
Every pH sensor needs to be matched to a specific instrument with buffer solutions. It is not possible to pre-calibrate a standard pH electrode, so it is not possible to have an accuracy value. To establish accuracy requires
the simultaneous use of the sensor, pH instrument and buffer solutions. Only when used simultaneously can
accuracy be determined. Therefore, any standard pH sensor by itself does not have a defined accuracy.
Structure of an ISM® pH measuring loop
Intelligent Sensor Management (ISM) is METTLER TOLEDO’s digital technology for in-line analytical sensors.
An ISM pH loop consists of a combination pH sensor with a built-in analog-to-digital converter and memory,
connected with a cable to a pH instrument which can receive the digital signal.
The analog part of the pH sensor is identical, as described earlier, to standard pH sensors. The difference is in
the sensor’s head which contains the analog-to-digital converter, and a microprocessor. The signal output by
ISM sensors is digital. This low impedance signal is very robust. Humidity, which can affect analog signals, is
not a problem and longer cables can be installed.
The ISM sensor’s microprocessor has the ability to record
and retain calibration data. This means that ISM sensors do
not need to be calibrated at the process. Instead, they can be
calibrated in any convenient location using an ISM transmitter
or METTLER TOLEDO’s iSense PC software. After calibration,
the pH sensor can be stored until it is required.
When a pre-calibrated ISM pH sensor is installed at a process,
the calibration data is automatically read by the connected ISM
transmitter, which adjusts itself without any operator intervention.
The pH buffer solutions used for calibration are traceable to
accepted standards (e.g. NIST) with a given accuracy and
therefore ISM pH sensors can be specified as sensors with a
defined accuracy.
Hints and Tips to Improve Accuracy
METTLER TOLEDO Process Analytics
7
Requirements for a high quality pH sensor
Electrode designed for bioprocesses
pH sensitive glass
Liquid junction (diaphragm)
Reference electrolyte
sterilizable, autoclavable,
minimal zero point shift after sterilization or CIP
minimal clogging during fermentation, easy to clean
free from silver ion, compatible with protein containing media
Error analysis for sensors, buffers and instruments
In error analysis the worst case in error calculations is achieved by building the sum of all possible errors in one
direction, and not allowing any compensation of errors.
pH buffers
pH electrode
Measuring value
7.20 pH
Diffusion potential
1 mV
Error by diff. pot.
-0.02 pH
Slope 250C 98%
Technical buffers
pH transmitter
Temperature sensor
Uncertainty meas.
Uncertainty current
Scaling 4 mA
Scaling 20 mA
Resolution
Current output
Uncertainty output
+/- 0.02 pH
+/- 0.050 mA
2 pH
12 pH
0.63 pH/mA
1.6000 mA/pH
12.320 mA
+/- 0.050 mA
+/- 0.03 pA
Worst case
8
+/- 0.02 pH
Hints and Tips to Improve Accuracy
METTLER TOLEDO Process Analytics
0.01 pH ISM Temperature 370 C
Error temp. meas.
-20 C
Slope 250 C, practical
-57.98 mV/pH
Slope 370 C, practical
-60.31 mV/pH
Slope with temp. error -59.92 mV/pH
Temperature error
+/- 0.00pH
0.01 pH
ISM digital communication
Uncertainty pH electrode
Uncertainty buffer
Uncertainty temperature
Uncertainty transmitter
Uncertainty current output
Uncertainty total
+/- 0.02pH
+/- 0.02pH
+/- 0.00pH
+/- 0.02pH
+/- 0.03pH
+/- 0.09pH
+/- 0.04 with ISM
With ISM pH transmitters the uncertainty of measurement and the uncertainty of output is eliminated, and as
can be seen above, the maximum error of +/- 0.09 pH can be reduced to +/- 0.04 pH.
5. Accuracy of dissolved oxygen sensors
Amperometric oxygen sensors
The measuring principle of such sensors is as follows. Oxygen diffuses through a
semi-permeable membrane and is reduced at a cathode which is held at a defined
potential. The reduction current is proportional to the partial pressure of oxygen in the
media.
I = K x A x D x S x 1/d x pO2
The relationship between sensor current “I” and partial pressure is only valid if
membrane thickness remains constant. During sterilization, the membrane is slightly
stretched, resulting in a higher sensor output. This effect reduces after each sterilization cycle. Small handling errors during membrane replacement, such as not removing
electrolyte drops on the membrane module shaft, can lead to a small membrane
blow-out. This can be very critical in terms of accuracy as membrane thickness has
been altered.
Linearity of sensor signal pO2 < 0.21 bar +/- 1%
Amperometric DO
sensor with membrane
module (cut view)
Stability of residual current and slope (air current) typically < 2%/week
Optical dissolved oxygen sensors
Optical oxygen technology is based on fluorescence quenching. In contrast to the amperometric sensor, which
detects the reduction current of oxygen, the optical method measures energy transfer between a fluorescing
chromophore (fluorophore) and oxygen.
For further information, download our “Good Operating Procedures for Optical Dissolved Oxygen Sensors”
www.mt.com/Optical-GoP-Guide
Optical sensors do not contain any electrolyte or replaceable membrane. In METTLER TOLEDO’s InPro 6860i,
the lifetime of the oxygen-sensing element, the OptoCap, is dependent on the quantity of light it is exposed to.
The OptoCap is the only part of the sensor which must be replaced periodically. Typically, in fermentation applications a lifetime exceeding 6 months can be expected.
Hints and Tips to Improve Accuracy
METTLER TOLEDO Process Analytics
9
InPro 6860i optical DO sensor with OptoCap
relaceable oxygen-sensing element
InPro 6860i optical
DO sensor
Due to its measurement technology, handling errors such as seen with amperometric sensors are avoided with
optical sensors. Further, the maintenance requirement is significantly lower than with amperometric sensors.
Another substantial advantage of the design of the InPro 6860i is very low measurement drift: four to five times
lower than typical amperometric sensors (< 0.5%/week).
Stability control for optical DO sensors
The degradation of the fluorophore is very linear over the lifetime of an OptoCap. Occasional calibration corrects
for measurement drift observed in normal sensor ageing. The InPro 6860i utilizes a proprietary alogorithm
that monitors the sample rate, oxygen level and process temperature to accurately compensate for shifts in Phi
values caused by fluorophore degradation. This so-called stability control stabilizes the oxygen reading and
significantly reduces the need for frequent calibration.
The stability control algorithm is able to learn process specific sensor ageing. By performing a one-point calibration in air after a few batches, the sensor compares calculated Phi shift values with calibration Phi shift to
accurately compensate for future fluorophore degradation. As a result, sensor drift is minimal. This represents a
major benefit for long duration batches with cell cultures.
10
Hints and Tips to Improve Accuracy
METTLER TOLEDO Process Analytics
Calibration
The most common calibration can be easily performed by holding the sensor in air. In fermentation or cell
cultures, calibration is done after sterilization in the air saturated media. This can be applied to amperometric or
optical sensors.
2-point calibration
Many pharmaceutical companies dictate a 2-point calibration in their SOPs. Additionally to the high oxygen
calibration (air saturation) a zero point calibration must be performed.
For optical DO sensors a 2-point calibration is required with every OptoCap exchange. For correct calibration,
nitrogen gas or other oxygen-free medium with a purity level of at least 99.99% should be used to achieve the
Phi 0 point, followed by exposure to air for the Phi 100 point. 2-point calibrations are performed on METTLER
TOLEDO transmitters or on a computer running iSense software. The resulting calibration curve is stored on the
sensor and is continuously referenced as the sensor makes its measurements.
For normal use a zero point calibration is not necessary. It should be noted that a faulty zero calibration with a
non-defined, oxygen-free medium will have a negative effect on accuracy. Some operators perform a zero-point
calibration during sterilization with the assumption that the bioreactor is oxygen free. Such a procedure is not
recommended as the measuring temperature is very different to the specified range. This will result in poor
measurement accuracy.
1-point calibration
A 1-point calibration establishes the slope (for amperometric sensors) or a new Phi 100 value (for optical
sensors). 1-point calibrations can calibrate sensors in air with measurement settings at 100% and local atmospheric pressure.
Process calibration
Process calibrations differ from 1-point calibration in that the former are conducted with a sensor in situ in a
reactor. For optical sensors, process calibration establishes a new Phi100 value against the stored calibration
curve to create a new curve. Because calibration curves for optical sensors are not linear, 1–point process
calibrations in vessels with head-space must accurately account for system pressures or risk jeopardizing the
accuracy of the entire curve. Because of this, we recommend using 1-point process scaling instead of process
calibration for the vast majority of post-SIP applications.
Process scaling
Unlike 1-point process calibrations, process scaling sets the measurement value to a desired level without
making any adjustments to the calibration curve. With this method the real process pressure and the solubility
factor for oxygen can be ignored, resulting in improved accuracy.
Hints and Tips to Improve Accuracy
METTLER TOLEDO Process Analytics
11
6. Conclusions
In bioprocess applications, maintaining a culture in an acceptable range around the optimal setpoints is key for
efficient and consistent production and minimal byproduct production.
To achieve the best possible accuracy from in-line pH and DO sensors, the following points are important:
1. Choose sensors which are designed for bioprocessing and SIP/CIP conditions. For pH, select sensors with
A41 membrane glass and pre-pressurized liquid reference electrolyte. For dissolved oxygen sensors we recommend optical sensors.
2. Sensor must be cleaned prior to calibration.
3. Calibration is essential for ensuring measurement accuracy. Consistently follow approved calibration procedures. For pH sensors always use fresh buffers. For DO sensors perform zero calibration only with defined
non-oxygen media.
4. Measure pH in the lab at the same temperature as in the process for process calibration.
5. Errors due to handling or faulty buffers directly lead to low accuracy.
Intelligent Sensor Management (ISM) can improve sensor
accuracy
1. ISM’s digital signal is very robust and many errors due to analog signal transmission and can be eliminated.
2. Calibration without handling errors can be performed off-line with iSense software in any convenient location
rather than in the production area.
3. ISM optical dissolved oxygen sensors feature very low maintenance. Less handling will improve accuracy.
4. Pre-batch diagnostics ensure use of sensors that will operate reliably throughout a batch.
12
Hints and Tips to Improve Accuracy
METTLER TOLEDO Process Analytics
Best Practice Guides
Maintaining product quality while improving process safety and controlling operating costs is a constant challenge in the process industries.
METTLER TOLEDO Process Analytics has created several Best Practice Guides for different industries with examples of how advanced process analytical systems can help operators increase product quality and yield, while
at the same time reduce operating costs.
Find the appropriate guide for your specific needs and download your free copy today!
Get access here: www.mt.com/pro-guides
pH and DO in Theory and Practice
pH Theory Guide: A guide to pH
Measurement – the Theory and Practice of
pH Applications
DO Guide: Good Operating Procedures for
Optical Dissolved Oxygen Sensors
Pharmaceutical Industry
Pharma Guide “Achieving the Highest
Level of Performance in Bioreactor Process
Control”
Pharmaceutical Waters Guide for
Regulatory Compliance, Analysis and Realtime Release
Hints and Tips to Improve Accuracy
METTLER TOLEDO Process Analytics
13
In-line Analytics Website
Dedicated to the Pharmaceutical Industry
The METTLER TOLEDO Process Analytics website for the
pharmaceutical industry is packed with information on how
our in-line measurement solutions improve process reliability,
increase production yield and reduce operating costs.
Visit our website to:
• Discover our extensive portfolio of sensors and transmitters.
• Download white papers and application notes, watch videos
and webinars.
• Find out how intelligent measurement solutions can prevent
batch losses and simplify sensor documentation.
www.mt.com/ism-pharma
www.mt.com/ism-pharma
www.mt.com/pro
For more information
METTLER TOLEDO Process Analytics
Im Hackacker 15
8902 Urdorf
Switzerland
http://www.mt.com/pro
Subject to technical changes
© 02/2015 METTLER TOLEDO Process Analytics