User’s
Manual
TDLS8000
Tunable Diode Laser Spectrometer
IM 11Y01D01-01EN
IM 11Y01D01-01EN
3rd Edition
i
u Introduction
Thank you for purchasing the TDLS8000 Tunable Diode Laser Spectrometer.
Please read the following respective documents before installing and using the TDLS8000.
The description of following products also be included in this manual.
YH8000
IF8000
YC8000
HMI Interface Unit
Isolation Flange for TDLS8000
Flow Cell for TDLS8000
The related documents are as follows.
General Specifications GS 11Y01D01-01EN
User’s Manual
IM 11Y01D01-01EN (this manual)
* the “EN” in the document number is the language code.
n Notes on Handling User’s Manuals
• Please hand over the user’s manuals to your end users so that they can keep the user’s
manuals on hand for convenient reference.
• Please read the information thoroughly before using the product.
• The purpose of these user’s manuals is not to warrant that the product is well suited to any
particular purpose but rather to describe the functional details of the product.
• No part of the user’s manuals may be transferred or reproduced without prior written
consent from YOKOGAWA.
• YOKOGAWA reserves the right to make improvements in the user’s manuals and product at
any time, without notice or obligation.
• If you have any questions, or you find mistakes or omissions in the user’s manuals, please
contact our sales representative or your local distributor.
n Drawing Conventions
Some drawings may be partially emphasized, simplified, or omitted, for the convenience of
description.
Some screen images depicted in the user’s manual may have different display positions or
character types (e.g., the upper / lower case). Also note that some of the images contained in this
user’s manual are display examples.
n Notes on Hardware
l Appearance and Accessories
Check the following when you receive the product:
• Appearance
• Standard accessories
l Model and Suffix Codes
The name plate on the product contains the model and suffix codes. Compare them with those in
the general specification to make sure the product is the correct one. If you have any questions,
contact our sales representative or your local distributor.
Media No. IM 11Y01D01-01EN
3rd Edition : Jun. 2016 (YK)
All Rights Reserved Copyright © 2015, Yokogawa Electric Corporation
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
ii
u
Safety Precautions
n Safety, Protection, and Modification of the Product
• In order to protect the system controlled by the product and the product itself and ensure
safe operation, observe the safety precautions described in this user’s manual. We assume
no liability for safety if users fail to observe these instructions when operating the product.
• If TDLS8000 and YH8000 are used in a manner not specified in this user’s manual, the
protection provided by these instruments may be impaired.
• If any protection or safety circuit is required for the system controlled by the product or for
the product itself, prepare it separately.
• Be sure to use the spare parts approved by Yokogawa Electric Corporation (hereafter
simply referred to as YOKOGAWA) when replacing parts or consumables.
• Modification of the product is strictly prohibited.
• The following safety symbols are used on the product as well as in this manual.
WARNING
This symbol indicates that an operator must follow the instructions laid out in this manual in order
to avoid the risks, for the human body, of injury, electric shock, or fatalities. The manual describes
what special care the operator must take to avoid such risks.
CAUTION
This symbol indicates that the operator must refer to the instructions in this manual in order to
prevent the instrument (hardware) or software from being damaged, or a system failure from
occurring.
CAUTION
This symbol gives information essential for understanding the operations and functions.
NOTE
This symbol indicates information that complements the present topic.
This symbol indicates Protective Ground Terminal.
This symbol indicates Function Ground Terminal. Do not use this terminal as the protective
ground terminal.
n Warning and Disclaimer
The product is provided on an “as is” basis. YOKOGAWA shall have neither liability nor
responsibility to any person or entity with respect to any direct or indirect loss or damage arising
from using the product or any defect of the product that YOKOGAWA can not predict in advance.
n Safety Precautions for Explosion Protected Type Instrument
Specified types of TDLS8000 and YH8000 are designed to protect against explosion.
When these type instruments are used in a hazardous area, please be sure to read Appendix 8.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
iii
CAUTION
Only trained persons use TDLS8000 and YH8000 in industrial locations.
n TDLS8000 and YH8000
There are safety symbols in the point of the figure to a product.
TDLS8000
YH8000
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
iv
CAUTION
Be careful not to connect the power supply wires to the incorrect locations or reverse the polarity.
CAUTION
Use cables with a durable temperature of at least 70 °C.
• Don’t install “general purpose type” instruments in the hazardous area.
• The Instrument is packed carefully with shock absorbing materials, nevertheless, the
instrument may be damaged or broken if subjected to strong shock, such as if the
instrument is dropped. Handle with care.
• Components that can be damaged by static electricity are used in the TDLS8000 tunable
diode laser spectrometer and YH8000 HMI interface unit. Take protective measures against
static electricity when performing maintenance and inspection and use conductive packing
material for shipping replacement components.
• Do not use an abrasive or organic solvent for cleaning the TDLS8000 tunable diode laser
spectrometer and YH8000 HMI interface unit.
• TDLS8000 and YH80000 are EN61326-1 Class A products, and it is designed for use in the
industrial environment. Please use these instruments in the industrial environment only.
• The HART communication may be influenced by strong electromagnetic field.
In this case another trial of the HART communication and/or operation with TDLS8000 touch
screen can be carried out.
WARNING
Depending on the specifications, toxic CO and NH3 gas may be used for the offline calibration of
this product. Take special care and ensure correct use when using such gas.
WARNING
Sufficiently ventilate the room to ensure the purge gas does not accumulate and there is no
shortage of oxygen.
CAUTION
Do not subject the equipment to an impact. It may cause irreparable damage to the laser.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
v
CAUTION
Sufficiently understand this user’s manual and carry out the work carefully so as not to make a
mistake with a pipe or wire.
CAUTION
Electrostatic discharge
The TDLS8000 and YH8000 contains devices that can be damaged by electrostatic discharge.
When servicing this equipment, please observe proper procedures to prevent such damage.
Replacement components should be shipped in conductive packaging. Repair work should
be done at grounded workstations using grounded soldering irons and wrist straps to avoid
electrostatic discharge.
CAUTION
• Do not use an abrasive or organic solvent in cleaning the instrument.
CAUTION
Please turn off the power to the TDLS8000 befor remove the analyzer from process flange.
n Maintenance by qualified engineer
Work carried out by other than a qualified engineer may cause injury to the worker and/or severe
damage to the equipment. Furthermore, if the warnings in this manual are not observed, the
worker may be seriously injured and/or the equipment may be severely damaged.
Maintenance of the equipment must be performed by a qualified engineer. Qualified engineer
refers to the following:
• Engineer who is familiar with how to safely handle process analyzers (or general automation
technology) and has read this manual and understood its content.
• Engineer who has received training on how to start and configure equipment and has read
this manual and understood its content.
n Replacement of battery
The battery (CR2050 type) on the CPU board in TDLS8000 cannot be installed on site because it
must be mounted at the factory. If it needs replacing, contact a Yokogawa service center.
n Transportation of products containing lithium batteries
TDLS8000 contains lithium batteries. Primary lithium batteries are regulated in transportation by
the U.S. Department of Transportation, and are also covered by the International Air Transport
Association (IATA), the International Civil Aviation Organization (ICAO), and the European
Ground Transportation of Dangerous Goods (ARD). It is the responsibility of the shipper to
ensure compliance with these or any other local requirements. Please consult the current
regulations and requirements regarding the transportation of lithium batteries before shipping.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
vi
n How to dispose the batteries
This is an explanation about the new EU Battery Directive(DIRECTIVE 2006/66/EC).This
directive is only valid in the EU.
Batteries are included in TDLS8000. Batteries incorporated into this product cannot be removed
by yourself. Dispose them together with TDLS8000.
When you dispose TDLS8000 in the EU, contact your local Yokogawa Europe B.V.office. Do not
dispose them as domestic household waste.
Battery type: Manganese dioxide lithium battery
Notice:
The symbol means they shall be sorted out and collected as ordained in
ANNEX II in DIRECTIVE 2006/66/EC.
n Product Disposal
The instrument should be disposed of in accordance with local and national legislation/regulations.
n Safety Precautions for Laser Products
TDLS8000 uses a laser light source. TDLS8000 is a Class 1 laser product as defined by
IEC60825-1 Safety of Laser Products—Part1: Equipment Classification, Requirements and
User’s Guide. In addition, TDLS8000 complies with 21 CFR 1040.10 and 1040.11 except for
deviations pursuant to Laser Notice No. 50, dated June 24, 2007.
Complies with 21 CFR 1040.10 and 1040.11 except for deviations
pursuant to Laser Notice No.50, dated June 24, 2007
2-9-32 Nakacho, Musashino-shi, Tokyo, 180-8750 Japan
Laser Unit
Sensor Control Unit
CAUTION
This analyzer, a class 1 invisible laser product, is safe enough to avoid eye injury although,
please do not see a light source. Laser light is emitted from the laser unit right after an analyzer is
powered on. After attaching both the laser unit and sensor control unit to a process flange or flow
cell, please power on an analyzer while laser light is not emitted outside measurement process.
n Safety and EMC conforming standards
With regard to standards of Explosion Protect, please see Appendix 8.
l TDLS8000 Tunable Diode Laser Spectrometer
Safety Conforming Standards:
CE
EN61010-1, EN61010-2-030
UL
UL61010-1, UL 61010-2-030
CSA
CAN/CSA-C22.2 No.61010-1, CAN/CSA-C22.2 No.61010-2-030
GB
GB30439 Part 1
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
vii
Installation altitude: 2000 m or less
Installation category: I (Anticipated transient overvoltage 330V)
Measuring category: O (Other)
Pollution degree:
2, Indoor/Outdoor use
Note: Installation category, called overvoltage category, specifies impulse withstand voltage. Pollution degree indicates the
degree of existence of solid, liquid, gas or other inclusions which may reduce dielectric strength.
EMC Conforming Standards:
CE
EN55011 Class A Group 1
EN61326-1 Class A Table 2 (For use in industrial location), EN61326-2-3
RCM
EN55011 Class A Group 1
KC
KN11 Class A Group 1, KN61000-6-2 (Korea Electromagnetic Conformity)
한국 전자파적합성 기준
A급 기기 (업무용 방송통신기자재)
이 기기는 업무용(A급) 전자파적합기기로서 판매자 또는
사용자는 이 점을 주의하시기 바라며, 가정외의 지역에서
사용하는 것을 목적으로 합니다.
Cable conditions:
• Power cable Use a shielded cable.
• I/O cable
Use a shielded cable.
• Cable for connecting between units
Use a separately sold dedicated cable (shielded cable). Attach the ferrite clamp
supplied with TDLS8000 to the cable connecting between units on the inside of
the equipment in the case of both an SCU and LU.
• Ethernet cable Use an STP cable (shielded) of category 5e or higher.
Influence in immunity environment (criteria A):
Variation of gas concentration measurement value
Within ±15% of reading
l YH8000 HMI Unit
Safety Conforming Standards:
CE EN61010-1
UL UL61010-1
CSA
CAN/CSA-C22.2 No.61010-1
GB
GB30439 Part 1
Installation Altitude: 2000 m or less
Installation category: I (Anticipated transient overvoltage 330 V)
Pollution degree:
2, Indoor/Outdoor use
EMC Conformity Standards:
CE
EN55011 Class A Group 1
EN61326-1 Class A Table 2 (For use in industrial location)
RCM
EN55011 Class A Group 1
KC
KN11 Class A Group 1, KN61000-6-2 (Korea Electromagnetic Conformity)
한국 전자파적합성 기준
A급 기기 (업무용 방송통신기자재)
이 기기는 업무용(A급) 전자파적합기기로서 판매자 또는
사용자는 이 점을 주의하시기 바라며, 가정외의 지역에서
사용하는 것을 목적으로 합니다.
Cable conditions:
• Power cable Use a shielded cable.
• Local HMI connection cable
Use a separately sold dedicated cable (shield cable).
• Ethernet cable Use an STP cable (shielded) of category 5e or higher.
Product conformity assessments of YH8000 for the relevant standards are performed in its own
right.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
viii
n ATEX Documentation
The procedure is only applicable to the countries
in European Union.
GB
All instruction manuals for ATEX Ex related
products are available in English, German and
French. Should you require Ex related instructions
in your local language, you are to contact your
nearest Yokogawa office or representative.
DK
Alle brugervejledninger for produkter relateret
til ATEX Ex er tilgængelige på engelsk, tysk og
fransk. Skulle De ønske yderligere oplysninger
om håndtering af Ex produkter på eget sprog, kan
De rette henvendelse herom til den nærmeste
Yokogawa afdeling eller forhandler.
I
Tutti i manuali operativi di prodotti ATEX
contrassegnati con Ex sono disponibili in inglese,
tedesco e francese. Se si desidera ricevere i
manuali operativi di prodotti Ex in lingua locale,
mettersi in contatto con l’ufficio Yokogawa più
vicino o con un rappresentante.
E
Todos los manuales de instrucciones para
los productos antiexplosivos de ATEX están
disponibles en inglés, alemán y francés. Si desea
solicitar las instrucciones de estos artículos
antiexplosivos en su idioma local, deberá ponerse
en contacto con la oficina o el representante de
Yokogawa más cercano.
NL
Alle handleidingen voor producten die te maken
hebben met ATEX explosiebeveiliging (Ex)
zijn verkrijgbaar in het Engels, Duits en Frans.
Neem, indien u aanwijzingen op het gebied van
explosiebeveiliging nodig hebt in uw eigen taal,
contact op met de dichtstbijzijnde vestiging van
Yokogawa of met een vertegenwoordiger.
SF
Kaikkien ATEX Ex -tyyppisten tuotteiden
käyttöhjeet ovat saatavilla englannin-, saksan- ja
ranskankielisinä. Mikäli tarvitsette Ex -tyyppisten
tuotteiden ohjeita omalla paikallisella kielellännne,
ottakaa yhteyttä lähimpään Yokogawa-toimistoon
tai -edustajaan.
P
Todos os manuais de instruções referentes
aos produtos Ex da ATEX estão disponíveis
em Inglês, Alemão e Francês. Se necessitar
de instruções na sua língua relacionadas
com produtos Ex, deverá entrar em contacto
com a delegação mais próxima ou com um
representante da Yokogawa.
F
Tous les manuels d’instruction des produits
ATEX Ex sont disponibles en langue anglaise,
allemande et française. Si vous nécessitez des
instructions relatives aux produits Ex dans votre
langue, veuillez bien contacter votre représentant
Yokogawa le plus proche.
D
Alle Betriebsanleitungen für ATEX Ex bezogene
Produkte stehen in den Sprachen Englisch,
Deutsch und Französisch zur Verfügung. Sollten
Sie die Betriebsanleitungen für Ex-Produkte in
Ihrer Landessprache benötigen, setzen Sie sich
bitte mit Ihrem örtlichen Yokogawa-Vertreter in
Verbindung.
S
Alla instruktionsböcker för ATEX Ex
(explosionssäkra) produkter är tillgängliga på
engelska, tyska och franska. Om Ni behöver
instruktioner för dessa explosionssäkra produkter
på annat språk, skall Ni kontakta närmaste
Yokogawakontor eller representant.
GR
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
ix
SK
PL
CZ
SLO
LT
H
BG
LV
EST
RO
M
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
x
n Authorised Representative in EEA
The Authorised Representative for TDLS8000 and YH8000 in EEA is Yokogawa Europe B.V.
(Euroweg 2, 3825 HD Amersfoort, The Netherlands).
n Trademark Notices
• Ethernet is a registered trademark of XEROX Corporation.
• Modbus are registered trademarks of Schneider Electric SA.
• All other company and product names mentioned in this user’s manual are trademarks or
registered trademarks of their respective companies.
• We do not use TM or ® mark to indicate those trademarks or registered trademarks in this
user’s manual.
n Partial change to the alarm specifications
If you did not purchase the TDLS8000 and YH8000 at the same time, the software versions may
be different. Note the following.
A portion of the alarm specifications was changed in software version 1.02.01.
Displaying alarms correctly on the YH8000 when the TDLS8000 and YH8000 are used in
combination requires the YH8000 software version to also be 1.02.01 or later.
YH8000
1.01.xx
1.02.xx
Can be used with A portion of the
1.01.xx the previous alarm alarms will not be
specifications.
displayed correctly.
TDLS8000
A portion of the
Can be used with
1.02.xx alarms will not be the new alarm
displayed correctly. specifications.
If the software version of either device is old, update the old version to the new version. For
details on software updating, contact your nearest Yokogawa representative.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
Toc-1
TDLS8000
Tunable Diode Laser Spectrometer
IM 11Y01D01-01EN 3rd Edition
CONTENTS
u Introduction.....................................................................................................i
u
Safety Precautions........................................................................................ii
1. Overview..................................................................................................... 1-1
2.
3.
1.1
System configuration........................................................................................ 1-1
1.2
Name and Function of Each Part..................................................................... 1-3
Specifications............................................................................................ 2-1
2.1
TDLS8000 Tunable Diode Laser Spectrometer Specifications.................... 2-1
2.2
Others.................................................................................................................. 2-6
2.2.1
YH8000 HMI Unit................................................................................ 2-6
2.2.2
IF8000 Isolation Flanges.................................................................... 2-7
2.2.3
YC8000 Flow Cell............................................................................... 2-8
2.2.4
Calibration Cell.................................................................................... 2-9
2.2.5
Unit Connection Cable........................................................................ 2-9
2.3
Model and Codes............................................................................................. 2-10
2.4
External Dimensions....................................................................................... 2-13
Installation, Wiring, Optical Axis Adjustment, and Piping.................... 3-1
3.1 Installation.......................................................................................................... 3-1
3.1.1
Measurement Point Selection............................................................. 3-2
3.1.2
Constructing Process Flanges............................................................ 3-3
3.1.3
Attaching the TDLS8000 to the Process Flange................................ 3-5
3.2 Wiring................................................................................................................ 3-10
3.3
3.2.1
Connecting between the Sensor Control Unit (SCU) and Laser Unit (LU).. 3-14
3.2.2
Connecting the Power Cable and Grounding................................... 3-15
3.2.3
Connecting to Temperature and Pressure Transmitters................... 3-16
3.2.4
Wiring Analog Outputs (AO)............................................................. 3-18
3.2.5
Wiring Digital Outputs....................................................................... 3-18
3.2.6
Wiring Digital Inputs.......................................................................... 3-20
3.2.7
Wiring Solenoid Valve Control Outputs............................................ 3-20
3.2.8
Connecting an Ethernet Cable......................................................... 3-21
Optical Axis Adjustment.................................................................................. 3-23
3.3.1
Optical Axis Adjustment When Not Using an LAO Unit (Optical path length
6 m or less)........................................................................................ 3-24
3.3.2
Optical Axis Adjustment When Using an LAO Unit (Optical path length 6
m or more)......................................................................................... 3-25
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
Toc-2
3.4 Piping................................................................................................................ 3-29
4.
3.4.1
Purge Gas Piping for In-situ Installation........................................... 3-31
3.4.2
Purge Gas Piping for Sampling System Using Flow Cells............... 3-32
3.4.3
Purge Gas Piping for Explosionproof/Flameproof type.................... 3-33
YH8000 Installation................................................................................... 4-1
4.1
Local HMI Installation........................................................................................ 4-1
4.2
Wiring for Local HMI Installation...................................................................... 4-4
4.3
Remote HMI Installation.................................................................................... 4-7
4.4
Wiring for Remote HMI Installation.................................................................. 4-9
5. Startup........................................................................................................ 5-1
5.1
5.2
5.3
5.4
6.
Connecting the HART Configuration Tool...................................................... 5-1
5.1.1
Installing a DD File.............................................................................. 5-1
5.1.2
Connection Procedure........................................................................ 5-1
5.1.3
Basic Menu Configuration.................................................................. 5-2
Connecting to the YH8000................................................................................ 5-3
5.2.1
Initialization and Connection Procedure............................................. 5-3
5.2.2
Setting the IP Address........................................................................ 5-4
5.2.3
Connecting to the TDLS8000............................................................. 5-7
5.2.4
Handling Connection Failures.......................................................... 5-10
5.2.5
Basic Screen Configuration.............................................................. 5-10
Setting Basic Parameters............................................................................... 5-13
5.3.1
Setting the Date and Time................................................................ 5-14
5.3.2
Setting the Process Optical Path Length.......................................... 5-17
5.3.3
Setting the Process Pressure........................................................... 5-18
5.3.4
Setting the Process Temperature..................................................... 5-18
5.3.5
Setting the Output Range................................................................. 5-19
5.3.6
Setting Process Alarms..................................................................... 5-20
Loop Check (Simulation output).................................................................... 5-23
5.4.1
Executing a Loop Check................................................................... 5-23
5.4.2
Auto Release Function..................................................................... 5-23
Configuration............................................................................................. 6-1
6.1
Process Parameter Settings............................................................................. 6-1
6.1.1
Process Optical Path Length.............................................................. 6-1
6.1.2
Process Pressure............................................................................... 6-1
6.1.3
Process Temperature.......................................................................... 6-2
6.2
Unit Settings....................................................................................................... 6-3
6.3
Analog Input Settings........................................................................................ 6-3
6.4
Analog Output Settings..................................................................................... 6-4
6.5
6.4.1
Normal Range Output......................................................................... 6-4
6.4.2
Output Hold......................................................................................... 6-4
Digital Output Settings...................................................................................... 6-6
6.5.1
DO Contact (DO-1)............................................................................. 6-6
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
Toc-3
6.5.2
6.6
Process Alarm Settings..................................................................................... 6-7
6.7
Digital Input Settings......................................................................................... 6-7
6.8
Valve Stream Settings....................................................................................... 6-8
6.9
6.10
6.11
7.
Fault Contact (DO-2)........................................................................... 6-7
6.8.1
Definitions of Stream Numbers........................................................... 6-8
6.8.2
Valve Usage Setting............................................................................ 6-9
6.8.3
Stream Settings................................................................................ 6-10
6.8.4
Initial Stream (Stream at Startup)..................................................... 6-11
Other Settings.................................................................................................. 6-12
6.9.1
Tag.................................................................................................... 6-12
6.9.2
Date and Time................................................................................... 6-12
6.9.3
User Password Setting..................................................................... 6-12
6.9.4
Display.............................................................................................. 6-13
6.9.5
Communication Address Setting...................................................... 6-14
6.9.6
Moving Average Count for Analysis Values...................................... 6-14
6.9.7
Concentration Offset......................................................................... 6-15
Non-process Parameter Settings................................................................... 6-15
6.10.1
Non-process Optical Path Length..................................................... 6-16
6.10.2
Non-process Pressure Setting.......................................................... 6-19
6.10.3
Non-process Temperature Setting.................................................... 6-19
6.10.4
Non-process Concentration Setting................................................. 6-19
Initializing the Settings (Factory Default Settings)...................................... 6-19
6.11.1
Initialization Procedure..................................................................... 6-19
6.11.2
Parameter Initial Value List............................................................... 6-20
HART Communication.............................................................................. 7-1
7.1 Connection......................................................................................................... 7-1
7.2
7.3
8.
Menu Tree........................................................................................................... 7-1
7.2.1
DD Menu............................................................................................. 7-1
7.2.2
DTM Menu (FieldMate)....................................................................... 7-5
Write Protection................................................................................................. 7-5
7.3.1
Hardware Write Protection.................................................................. 7-5
7.3.2
Software Write Protection................................................................... 7-6
7.3.3
Device Configuration Locked.............................................................. 7-8
7.4
Alarm Definition (Status group)....................................................................... 7-8
7.5
Functions Specific to HART Communication............................................... 7-10
7.5.1
Multidrop Mode................................................................................. 7-10
7.5.2
Squawk............................................................................................. 7-10
7.5.3
Aborting Calibration and Validation................................................... 7-10
YH8000 HMI Unit........................................................................................ 8-1
8.1 Connection......................................................................................................... 8-1
8.2
Home Screen...................................................................................................... 8-1
8.2.1
Home Screen Display Items............................................................... 8-2
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
Toc-4
8.3
Selecting the Style.............................................................................. 8-4
8.2.3
Setting the Meter Range..................................................................... 8-4
8.2.4
Alarm Indicator.................................................................................... 8-5
Trend Screen...................................................................................................... 8-6
8.3.1
Trend Screen Display Items................................................................ 8-7
8.3.2
Selecting the Items to Display............................................................. 8-8
8.3.3
Setting the Displayed Time................................................................. 8-9
8.3.4
Setting the Vertical Scale.................................................................. 8-10
8.4
Alarm Screen.................................................................................................... 8-10
8.5
Information Screen.......................................................................................... 8-11
8.6
8.7
9.
8.2.2
8.5.1
I/O List Screen.................................................................................. 8-12
8.5.2
Configuration View Screen............................................................... 8-12
8.5.3
System Information Screen.............................................................. 8-13
8.5.4
Spectrum Screen.............................................................................. 8-13
8.5.5
Alarm History Screen........................................................................ 8-14
8.5.6
Cal/Val History Screen...................................................................... 8-15
Configuration Screen...................................................................................... 8-16
8.6.1
TDLS8000 configuration screen....................................................... 8-16
8.6.2
YH8000 Configuration Screen.......................................................... 8-18
8.6.3
Setting the YH8000 Backlight........................................................... 8-19
When Multiple TDLS8000s Are Connected................................................... 8-20
8.7.1
Overall Display.................................................................................. 8-20
8.7.2
TDLS8000 Selection Screen............................................................ 8-21
8.7.3
Setting the Date and Time on the TDLS8000................................... 8-22
Inspection and Maintenance.................................................................... 9-1
9.1
9.2
9.3
9.4
Maintaining the Laser Beam and Transmission............................................. 9-1
9.1.1
Transmission Calibration.................................................................... 9-2
9.1.2
Process Window Cleaning.................................................................. 9-2
9.1.3
Insertion Tube Cleaning...................................................................... 9-5
Online Validation................................................................................................ 9-6
9.2.1
Preparation......................................................................................... 9-6
9.2.2
Configuration....................................................................................... 9-8
9.2.3
Execution............................................................................................ 9-9
9.2.4
Time Chart........................................................................................ 9-14
Mounting on a Calibration Cell....................................................................... 9-14
9.3.1
Preparation....................................................................................... 9-15
9.3.2
Preparation Procedure...................................................................... 9-16
9.3.3
Performing Calibration and Offline Validation................................... 9-19
9.3.4
Returning the TDLS8000 to the Process.......................................... 9-19
Offline Validation.............................................................................................. 9-21
9.4.1
Preparation....................................................................................... 9-21
9.4.2
Configuration..................................................................................... 9-23
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
Toc-5
9.5
9.6
10.
9.4.3
Execution.......................................................................................... 9-23
9.4.4
Time Chart........................................................................................ 9-24
Zero Calibration................................................................................................ 9-25
9.5.1
Preparation....................................................................................... 9-26
9.5.2
Configuration..................................................................................... 9-27
9.5.3
Execution.......................................................................................... 9-27
9.5.4
Time Chart........................................................................................ 9-28
Span Calibration.............................................................................................. 9-28
9.6.1
Preparation....................................................................................... 9-29
9.6.2
Configuration..................................................................................... 9-30
9.6.3
Execution.......................................................................................... 9-31
9.6.4
Time Chart........................................................................................ 9-32
9.7
Calibration Data Record and Restoring........................................................ 9-32
9.8
Automatic and Semi-automatic Execution of Validation and Calibration.9-33
9.8.1
Preparation....................................................................................... 9-34
9.8.2
Configuration..................................................................................... 9-34
9.8.3
Execution.......................................................................................... 9-35
9.8.4
Aborting the Stabilization Wait Time for Automatic or Semi-automatic
Execution.......................................................................................... 9-38
9.8.5
Consecutive Automatic Execution.................................................... 9-40
9.9
Analog Input Calibration................................................................................. 9-42
9.10
Analog Output Calibration.............................................................................. 9-42
9.11
Loop Check....................................................................................................... 9-43
9.12
Alarm History.................................................................................................... 9-43
Troubleshooting...................................................................................... 10-1
10.1
Fault Display and Handling............................................................................. 10-1
10.2
Warning Display and Handling....................................................................... 10-2
10.3
Handling Degraded Laser Transmission...................................................... 10-4
10.4
Process Window Replacement...................................................................... 10-7
10.4.1
Replacement Parts (Process window)............................................. 10-7
10.4.2
Process Window Replacement Procedure (Alignment flange)........ 10-8
10.4.3
Process Window Replacement Procedure (Process isolation flange)...10-9
10.4.4
Process Window Replacement Procedure (Flow cell)................... 10-10
10.5
Fuse Replacement.........................................................................................10-12
10.6
Communication Interruption during Manual Calibration and Validation.10-13
11. Modbus..................................................................................................... 11-1
11.1
Communication Specifications...................................................................... 11-1
11.1.1
Message Structure............................................................................ 11-1
11.1.2
Slave Response................................................................................ 11-2
11.2 Coil..................................................................................................................... 11-3
11.3
Input relay......................................................................................................... 11-4
11.4
Hold register..................................................................................................... 11-6
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
Toc-6
11.5
Input register.................................................................................................... 11-7
Appendix 1
Constructing Unit Connection Cables.............................. App.1-1
Appendix 2
Constructing Local HMI Connection Cables................... App.2-1
Appendix 3
General View of HART DD.................................................. App.3-1
Appendix 4
YH8000 Menu Tree.............................................................. App.4-1
Appendix 5
What Is an Analysis Period?.............................................. App.5-1
Appendix 6
Maintaining Good Transmission....................................... App.6-1
Appendix 7
Safety Instrumented System Installation......................... App.7-1
Appendix 8
Explosion Protected Type Instrument.............................. App.8-1
Customer Maintenance Parts List.................................... CMPL 11Y01D01-01EN
Customer Maintenance Parts List.................................... CMPL 11Y01D01-21EN
Customer Maintenance Parts List.................................... CMPL 11Y01D10-01EN
Customer Maintenance Parts List.....................................CMPL 11Y01D11-01EN
Customer Maintenance Parts List.................................... CMPL 11Y01D12-01EN
Revision Information................................................................................................i
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
1-1
<1. Overview>
1. Overview
Yokogawa’s new TDLS8000 houses all of the industry’s leading features in one robust device.
The platform design is for in situ measurements which negate the need for sample extraction and
conditioning.
The non-contacting sensor allows for a variety of process types including corrosive, abrasive and
condensing.
The first generation platform has been proven in many others for the measurements of O2, CO,
CH4, NH3, H2O and many more NIR absorbing gases.
This second generation platform has improved reliability and ease of installation and
maintenance while still meeting or exceeding designed application demands.
1.1
System configuration
Standard System Configuration
Laser unit (LU)
Sensor control unit (SCU)
Measured gas
Flow meter
Flow meter
24V DC±10%
Unit connection cable
Purge line for Optic
Purge line for Process window
Purge line for Process window
Purge line for Optic
System Configuration with YH8000 HMI Unit and Validation gas line
Laser unit (LU)
Sensor control unit (SCU)
Measured gas
Flow meter
Purge line for Optic
Flow meter
Unit connection cable
Purge line for Process window
Purge line for Process window
24V DC±10%
Purge line for Optic
Check gas line
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
1-2
<1. Overview>
Multi Analyzer Configuration with Remote HMI
Laser unit (LU)
Sensor control unit (SCU)
Measured
gas
24V DC
±10%
Measured
gas
24V DC±10%
Measured
gas
24V DC±10%
Measured
gas
Flow meter
Flow meter
Unit connection cable
24V DC±10%
Switching HUB
24V DC±10%
YH8000 HMI Unit
Note: If power supply is 100 to 240 V AC, purchase the Universal Power Supply M1276WW, separately.
If four multi analyzer configuration with remote HMI is made, five universal power supply including HMI are needed.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
1.2
1-3
<1. Overview>
Name and Function of Each Part
TDLS8000 is composed Laser Unit and Sensor Control Unit.
n TDLS8000 case
Nameplate
Display window
External earth terminal
Cover
*1:
Lock screw (*1)
Close the cover securely and fix it with a lock screw. Loosen the lock screw before opening the cover.
NOTE
Do not lose the lock screw when loosening it.
n TDLS8000 laser unit
LU display
Displays transmission.
Connecting terminal
LU fuse
Earth terminal for shield wires Jumper for the overvoltage
protection element
Usually leave it intact. (*1)
*1:
The TDLS8000 is equipped with an overvoltage protection element to prevent failure caused by surges and other overvoltages.
This element may hinder the correct measurement of the insulation resistance of the power line during insulation tests. To disable
this element, disconnect the jumper.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
1-4
<1. Overview>
l LU display
The four-digit LED displays the transmission value. The sequence of display after power-on is as
follows.
Timing
Description
For approx. 5 seconds No segment lights up.
after power-on
Until the first analysis
result is updated
Shown as
All segments light up.
Whenever the analysis The right-end dot blinks and
result is updated
transmission values are displayed.
(Example: 92.0 [%])
When the transmission is less than 1.0%, the value is displayed up to the second decimal place.
n TDLS8000 Sensor Control Unit
SCU display
USB port (*1)
HART write-protect switch
AI switch
Displays process statuses such as
gas concentration, transmission,
temperature, and pressure.
Switches for service (*2)
Ethernet connector
Used for connecting with
the YH8000 or
for Modbus communication
Open/close
knob
Connecting terminal
Main fuses
Fault LED (red)
Lights up when any
fault occurs.
DO LED (yellow)
Power LED (green) Lights up in conjunction with DO.
Lights up when the power is supplied.
Jumper for overvoltage
protection element
Usually leave it intact. (*3)
Earth terminal for shield wires
Connects the shield wire of connecting cables.
*1:
*2:
*3:
Service staff use the USB port for maintenance. Do not connect your USB devices.
Service staff use these switches for maintenance. Leave them all OFF.
The TDLS8000 is equipped with an overvoltage protection element to prevent failure caused by surges and other overvoltages.
This element may hinder the correct measurement of insulation resistance of the power line during insulation tests. To disable
this element, disconnect the jumper.
l SCU display
Starting screen
The screen below is displayed for approx. 10 seconds after power-on.
Current date and time
IP address
Software version
LCD starting screen
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
1-5
<1. Overview>
Normal screen
After the starting screen, the following screen is displayed. The presentation of concentration
values varies depending on the specifications of the TDLS8000.
1st line
2nd line
3rd line
4th line
5th line
6th line
HART communication icon
LCD normal screen (for single-gas measuring specifications)
HART communication icon
LCD normal screen (for double-gas measuring specifications)
NOTE
Measurements such as concentration and transmission are updated every analysis cycle. On the
SCU display, the temperature and pressure values are displayed alternately in the 4th line every
analysis cycle. This means that measurement is updated whenever the content of the 4th line
changes.
The details of the information displayed in each line are as follows.
Line
Item
1 Concentration of the 1st component gas (two lines are used for the singlegas measuring specifications) (*1)
2 Concentration of the 2nd component gas (for double-gas measuring
specifications) (*1)
3 Transmission
4 Temperature and pressure (displayed alternately for every analysis cycle)
1 Process pressure: Displays “pressure input mode pressure value”
Display example
O2 20.71%
CH4 1.82%
Trans 94.6%
PresAI1 101.32kPa
Pressure input mode
Display
Active Input:
Input source is AI-1.
Pres AI1
Input source is Modbus communication. Pres COM
Fixed
Pres Fix
2 Process temperature: Displays “temperature input mode temperature
value”
Temperature input mode
Active Input:
Input source is AI-2.
Input source is Modbus communication.
Fixed
Active Ambient
Active Peak (*2)
TempAI2 20.3ºC
Display
Temp AI2
Temp COM
Temp Fix
Temp ActA
Temp ActP
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
1-6
<1. Overview>
Line
Item
5 Status or alarm information
- Displays status information when there is no alarm.
- Displays alarm information when there is an alarm.
*: Displays the HART communication icon at the right end when HART
commands are received.
1 Status display: Displays the following equipment statuses.
Equipment status
During normal measurement
During warming-up
During maintenance
During calibration and
validation
During AO loop check or
calibration
During AI calibration
6
*1:
Measuring 12:10
Display example
Measuring hh:mm
Warm-up hh:mm
Maintenance hh:mm
Span Cal
(for span calibration)
AO1 Fixed=4.0mA
(for AO-1 4 mA output)
AI-1 (Pres) Cal
(for AI-1 calibration)
2 Alarm display: Displays “[W/F(alarm number)] alarm name”
- Alternates displays every 5 seconds when multiple alarms are
generated.
- [W##] means warning, [F##] means fault.
- Fault highlights characters.
Various setup information
Alternately displays the following items every 5 seconds.
1 IP address
6 HART address
3 LU temperature
4 SCU temperature
[F53] Trans Lost
IP: 192.168.1.10
HART ADRS: 0
LU: 34.5ºC
SCU: 33.4ºC
Displays the invalid value “***” while the following alarms are generated. Example: 02 ***%
Number
49
50
52
53
56
*2:
Display example
Alarm
Detector signal high
Peak center out of range
Absorption too high
Transmission lost
Outlier Rejection Limit
Can be set only for certain measurement target gas (application).
Spectrum screen
Absorption spectrum and receiving signals can be checked. Displaying the spectrum screen
requires changes in the setting of the TDLS8000. See “6.9.4 Display.”
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
1-7
<1. Overview>
n YH8000 HMI Unit
Ethernet connector (port 2)
Used for Modbus communication
Ethernet connector (port 1)
Used to connect with the TDLS8000
Connector
for service
(*3)
USB port (*1)
Power connecting terminal
Earth terminal
Screen upside-down switch
*1:
*2:
*3:
HMI fuse
Jumper for overvoltage protection element
Usually leave it intact. (*2)
Service staff use the USB port for maintenance. Do not connect your USB devices.
The YH8000 is equipped with an overvoltage protection element to prevent failure caused by surges and other overvoltages.
This element may hinder the correct measurement of insulation resistance of the power line during insulation tests. To disable
this element, disconnect the jumper.
Service staff use this connector for maintenance.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
Blank Page
2-1
<2. Specifications>
2.
Specifications
2.1
TDLS8000 Tunable Diode Laser
Spectrometer Specifications
Measurement object: O2, CO, CO or CH4, CO2, CO + CO2, H2O, NH3, NH3 + H2O, H2S, HCl
concentration in combustion exhaust gas and process gas
If other gas measurements are required, consult with Yokogawa.
Measurement system: Tunable diode laser spectroscopy
Light source; Near-infrared tunable diode laser
Measured components and ranges:
Measured component
O2
CO(ppm)
CO
CO or CH4 (*3)
CH4
NH3
H2O(ppm) in non HC (*1)
H2O(ppm) in HC (*2)
CO (%)
CO (%) + CO2 (%)
NH3
NH3 + H2O
H2O
H2S
CO2 (%) High Range
CO2 (%) Extend. Range
H2O (%)
HCl
*1:
*2:
*3:
Min. range
Max. range
0-1%
0-25%
0-200 ppm
0-10000 ppm
0-200 ppm
0-10000 ppm
0-5%
0-30 ppm
0-5000 ppm
0-30 ppm
0-30000 ppm
0-30 ppm
0-30000 ppm
0-20%
0-50%
0-30%
0-100%
0-30 ppm
0-5,000 ppm
0-5%
0-50%
0-5%
0-100%
0-1%
0-5%
0-30%
0-50%
0-10%
0-100%
0-50 ppm
0-5,000 ppm
Non hydrocarbon background
Hydrocarbon background
Please consult with Yokogawa if CO or CH4 ingredient coexists.
Please consult with Yokogawa if the measuring range for your sample gas is outside of the above
ranges
Optical path length:
Optical distance between the laser unit and the sensor control unit
Standard; 0.5 to 6 m (Application dependent)
Max;
30 m (With optional Large Aperture Optics (LAO))
Note: If your optical path length is under 0.5 m or over 30 m, please consult with Yokogawa.
Safety and EMC conforming standards:
Safety Conforming Standards:
CE
EN61010-1, EN61010-2-030
UL
UL61010-1, UL 61010-2-030
CSA
CAN/CSA-C22.2 No.61010-1, CAN/CSA-C22.2 No.61010-2-030
GB
GB30439 Part 1
Installation altitude: 2000 m or less
Installation category: I (Anticipated transient overvoltage 330V)
Measuring category: O (Other)
Pollution degree:
2, Indoor/Outdoor use
Note: Installation category, called overvoltage category, specifies impulse withstand voltage. Pollution degree indicates the degree of
existence of solid, liquid, gas or other inclusions which may reduce dielectric strength.
EMC Conforming Standards:
CE
EN55011 Class A Group 1
EN61326-1 Class A Table 2 (For use in industrial location), EN61326-2-3
RCM
EN55011 Class A Group 1
KC
KN11 Class A Group 1, KN61000-6-2 (Korea Electromagnetic Conformity)
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<2. Specifications>
2-2
Laser classification; CSA E60825-1-03(R2012), CE EN60825-1:2007, GB7247.1-2012
FDA 21CFR part 1040.10
Class 1 laser product
SIL Certification; IEC 61508:Functional safety of Electrical/electronic/programmable electronic
related systems; SIL 2 capability for single analyzer use, SIL 3 capability for
dual analyzer use.
Display:
128 x 64 dots LCD; On Sensor Control Unit
Status LEDs;
3 on Sensor Control Unit (Green: Power, Orange: DO, Red: Fault)
4 digit 7-segment LEDs; On Laser Unit
Display items: LCD on Sensor Control Unit; Gas concentration, Transmission, Process gas
temperature (AI), Process gas pressure (AI), System status, Alarm information,
System information (Product serial no., Laser module serial no., Output signal,
Setting parameter, IP address, HART address, Optical path length, Analyzer internal
temperature)
7-segment LEDs on Laser Unit;
Transmission
Analog output: 2 points, 4 to 20 mA DC (Isolated from the power supply and ground, Max. load
resistance 550 Ω)
Output types; Gas concentration, Transmission, Process gas temperature, Process gas
pressure
Output range; 3.0 to 21.6 mA DC
Digital communications:
HART; On analog output signal one (AO-1)
Load resistance;
250 to 550 Ω (Include cable resistance)
Ethernet; RJ-45 connector in Sensor Control Unit
Protocol; Modbus/TCP
Communication speed;
100 Mbps
Digital output: 2 points, contact rating 24V DC, 1A
DO;
Function: Activate during Warning / Calibration / Validation / Warm up / Maintenance
conditions
Contact Specification: Relay contact output (Isolated from the power supply and ground),
C-contact (NC/NO/COM)
Fault;
Function: Activate during normal conditions, not activate during Fault condition or when the
system power is off
Contact Specification: Relay contact output (Isolated from the power supply and ground),
A-contact (NC/COM)
Valve control output: 2 points
Function; Activate calibration or validation solenoid valves for zero, span or validation gas.
Output signal; 24V DC, 500 mA Max. per terminal
Alarm:
Warning; Gas concentration low, Gas concentration high, Transmission low, Process pressure
low, Process pressure high, Process temperature low, Process temperature high,
Validation required, Validation failure, Zero calibration error, Span calibration error,
Non process alarm, External alarm, Detector signal high, Absorption too high
Fault; Laser module temperature low, Laser module temperature high, Laser temperature
low, Laser temperature high, Peak center out of range, Reference peak height low,
Transmission lost, Reference transmission low, Reference peak height high, Laser
unit failure, Laser module error, File access error, E2PROM access error.
Digital input: 2 points
Function; External Alarm/Calibration start/Validation start/Stream switch (Valve control)
Contact specification; Zero voltage contact input (Isolated from the power supply and ground)
Input signal; Open signal: 100 kΩ or more, Close signal: 200 Ω or less
Analog input: 2 points
Signal type; 4 to 20 mA DC (Isolated from the power supply and Ground), with selectable
powered/unpowered function
Input signal range; 2.4 to 21.6 mA DC
Input types; Process gas temperature, Process gas pressure
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
2-3
<2. Specifications>
Transmitter power supply; 15 V DC or higher (at 20 mA DC)
26 V DC or less (at 0 mA DC)
Note: This voltage is generated between the AI terminals of TDLS8000. When calculating the minimum operating voltage for
transmitters, consider to allow margins for voltage drop in external wiring.
Self-diagnostics:
Laser Unit temperature, Sensor Control Unit temperature, Laser temperature, Detector
signal level, Memory read/write function, Peak locking condition
Calibration:
Calibration method; Zero/Span calibration
Calibration mode; Manual, Auto (Time initiate, Remote initiate (DI/Modbus), Semi-Auto
(YH8000/HART)
Validation:
Validation method; Up to 2 points
Validation mode; Manual, Auto (Time initiated, Remote initiate (DI/Modbus)), Semi-Auto
(YH8000/HART)
Power supply: 24V DC +/-10%
If your power supply is 100 to 240 V AC, Universal Power Supply, M1276WW
(sold separately), is required
Power consumption:
Max. 20W; TDLS8000 only
Max. 60W; TDLS8000 with YH8000 and 2 solenoid valves
Protection degree: IP66, Type 4X
Material: Case; Aluminum alloy
Wetted materials: 316 SS, BK-7 glass, Teflon encapsulated FKM (O-ring for alignment flange),
Silicone (O-ring for LAO)
Paint color: Mint green (RAL 190 30 15 or equivalent)
Weight:
Sensor Control Unit;
8 kg
Laser Unit; 8 kg
Large Aperture Optics;
14 kg
ANSI Class 150-2-RF (Eq.) Alignment Flange; 5 kg/pc
ANSI Class 150-3-RF (Eq.) Alignment Flange; 7 kg/pc
ANSI Class 150-4-RF (Eq.) Alignment Flange; 9 kg/pc
DIN PN16-DN50-D (Eq.) Alignment Flange; 5 kg/pc
DIN PN16-DN80-D (Eq.) Alignment Flange; 6 kg/pc
JIS 10K-50-FF (Eq.) Alignment Flange;
5 kg/pc
JIS 10K-80-FF (Eq.) Alignment Flange;
6 kg/pc
Flow Cell Alignment Flange;
5 kg/pc
Process gas condition:
Process gas temperature; Max. 1,500ºC, Application dependent
Process gas pressure;
Max.1 MPa abs., Min. 90 kPa abs., Application dependent
Max. 15 kPa G with LAO unit
Note: When using TDLS8000 as CE marking compliance product, it has following limitation.
General purpose model (-G1, -G2): The upper limit of the measurement gas pressure is 50kPa in gauge pressure.
ATEX model (-S1, -S2): The upper limit of the measurement gas pressure is 500kPa abs. The unstable gas
defined by following cannot be measured.
An unstable gas in this context is a gas liable to transform itself spontaneously, producing a sudden pressure increase.
Such transformation as an example can result from a relatively small variation of an operating parameter (e.g. pressure,
temperature, presence of catalyzing material) in a confined volume.
This includes gases that are classified as chemically unstable gases according to CLP Regulation (EC) No 1272/2008 as
amended.
Typical examples of unstable gases: acetylene (UN 1001), methyl acetylene (UN 1060), vinylfluoride (UN 1860), ozone and
dinitrogen oxide (UN 1067).
For further examples, see Table 35.1 of the UN Manual of Tests and Criteria.
Dust in process gas;
20 g/m3 or less
(Dust loading levels are dependent upon the application, OPL
and other installation factors)
Warm-up time: 5 min.
Installation condition:
Ambient operating temperature; -20 to 55ºC
Storage temperature;
-30 to 70ºC
Humidity;
0 to 95%RH at 40ºC (Non-condensing)
Mounting flange type;
ASME B16.5, DIN, JIS
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
2-4
<2. Specifications>
Cable entries;
Sensor Control Unit: 1/2NPT or M20x1.5mm,one hole
3/4NPT or M25x1.5mm, three holes
Laser Unit: 3/4 NPT or M25x1.5mm, one hole
Purge gas connections; 1/4NPT or Rc1/4
If other gas connections are required, please consult with Yokogawa.
Purge gas; Theoretically, instrument air could be used as a purge gas for all of the below
applications except for oxygen or H2O measurement. Choosing between using
nitrogen or instrument air or purge gas will ultimately depend upon further
application details and the desired precision of the measurement. All gasses should
be clean and dry.
Recommended purge gasses:
O2 analyzer:
N2 (99.99% or greater, application dependent)
H2O ppm analyzer: N2 (99.99% or greater with < 20 ppm H2O for feed to the optional dryer
package)
CO, CO or CH4, CO2, CO + CO2, NH3, NH3 + H2O, H2S, HCl analyzer:
N2 (99.99% or greater, application dependent) or Instrument air
Purge gas flow rates; 2 to 20 L/min for optic, 2 to 20 L/min and 150 mL/min for Div 1, Zone 1 Type
5 to 30 L/min for process window (Application dependent)
n Characteristics
Repeatability / Linearity:
Measured gas
O2
CO (ppm)
CO + CH4
CO
CH4
NH3
H2O (ppm) in non HC
H2O (ppm) in HC
CO (%)
CO
CO (%) + CO2 (%)
CO2
NH3
NH3 + H2O
H2O
H2S
CO2 (%) High Range
CO2 (%) Extend. Range
H2O (%)
HCl
Measurement conditions:
Repeatability
+/- 1% reading or +/- 0.01 %O2, whichever is greater
+/- 2% reading or +/- 1 ppm CO, whichever is greater
+/- 2% reading or +/- 1 ppm CO, whichever is greater
+/- 4% reading or +/- 0.02% CH4, whichever is greater
+/- 2% reading or +/- 1 ppm NH3, whichever is greater
+/- 2% reading or +/- 0.1 ppm H2O, whichever is greater
+/- 2% reading or +/- 0.1 ppm H2O, whichever is greater
+/- 1% reading or +/- 0.01% CO, whichever is greater
+/- 1% reading or +/- 0.1% CO, whichever is greater
+/- 1% reading or +/- 0.1% CO2, whichever is greater
+/- 2% reading or +/- 1 ppm NH3, whichever is greater
+/- 4% reading or +/- 0.05% H2O, whichever is greater
+/- 1% reading or +/- 0.005% H2S, whichever is greater
+/- 1% reading or +/- 0.005% CO2, whichever is greater
+/- 1% reading or +/- 0.02% CO2, whichever is greater
+/- 1% reading or +/- 0.004% H2O, whichever is greater
+/- 1% reading or +/- 2.5 ppm HCl, whichever is greater
Linearity
+/- 1% F.S.
+/- 1% F.S.
+/- 2% F.S.
+/- 4% F.S.
+/- 2% F.S.
+/- 1% F.S.
+/- 1% F.S.
+/- 1% F.S.
+/- 1% F.S.
+/- 1% F.S.
+/- 2% F.S.
+/- 2% F.S.
+/- 1% F.S.
+/- 1% F.S.
+/- 1% F.S.
+/- 1% F.S.
+/- 2% F.S.
Gas temperature; 25ºC, Gas pressure; 0.1 MPa, Optical path length; 1 m
Data Update Cycle:
Standard; Approx. 2 seconds (Response time may increase for non-standard
applications)
If less than 2 seconds response is required, please consult with Yokogawa
Zero Drift: Typically <0.1% of the minimum range over 24 months
Influences on the Measurement - Application dependent
A.
Temperature: The temperature of the measured gas should be taken into account by
the analyzer so that the reading can be corrected on a real time basis. The effect is
specific to each different measurement gas.
a.
If the gas temperature is constant at the desired measurement condition, then
a fixed gas temperature may be programmed into the analyzer. This fixed value
can be used in real time by the analyzer to provide a temperature-compensated
reading.
b.
If the gas temperature is relatively equal to the ambient temperature, then an
integral sensor value may be utilized by the analyzer. This active ambient value is
used real time by the analyzer to provide a temperature compensated reading.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<2. Specifications>
2-5
c.
If the gas temperature is variable, then an external sensor value may be utilized
by the analyzer. This active input value can be used in real time by the analyzer to
provide a temperature compensated reading.
B.
Pressure: The pressure of the measured gas must be taken into account by the
analyzer so that the reading can be corrected on a real time basis. The effect is
specific to each different measurement gas.
a.
If the gas pressure is constant at the desired measurement condition, then a fixed
gas pressure may be programmed to the analyzer. This fixed value can be used
in real time by the analyzer to provide a pressure compensated reading.
b.
If the gas pressure is variable, then an external sensor value may be utilized by
the analyzer. This active input value can be used in real time by the analyzer to
provide a pressure compensated reading.
n Hazardous area classifications
Division 1, Zone 1: Explosionproof
TDLS8000-D1 (FM Approval for US)
Division system:
Type of protection: Explosionproof for Class I, Division 1, Groups A, B, C, D, T5
Dust-Ignitionproof for Class II/III, Division 1, Groups E, F, G, T5
Enclosure rating: Type 4X
Applicable standards: FM Class 3600: 2011, FM Class 3615: 2006, FM Class 3616:
2011, FM Class 3810: 2005, NEMA 250: 2003
Zone system:
Type of protection: Class I, Zone 1, AEx d IIC T5 Zone 21, AEx tb IIIC T100ºC
Enclosure Rating: IP66
Applicable standards: ANSI/ISA-60079-0 2013, ANSI/ISA-60079-1 2009 (R2013),
ANSI/ISA-60079-31 2013, ANSI/IEC 60529 2004 (R2011)
TDLS8000-C1 (FM Approval for Canada)
Type of protection: Ex d IIC T5 Gb, Class II/III, Division 1, Groups E, F, G, T5
Enclosure rating:
IP66, Type 4X
Applicable standards: CAN/CSA-C22.2 No.0.4-04 (R2013), CAN/CSA-C22.2 No.0.5-1982 (R2012),
CAN/CSA-C22.2 No.25-1966 (R2014), CAN/CSA-C22.2 No.94.2-15,
CAN/CSA-C22.2 No.60079-0: 11, CAN/CSA-C22.2 No.60079-1: 11,
CAN/CSA-C22.2 No.60079-31: 12, CAN/CSA-C22.2 No.61010-1-12,
CAN/CSA-C22.2 No.60529-05 (R2010), ANSI/ISA-12.27.01-2011
TDLS8000-S1 (ATEX)
Type of protection: II 2(1) G Ex d [op is T6 Ga] IIC T5 Gc
II 2 D Ex tb IIIC T100ºC Db
Enclosure rating:
IP66 (In accordance with EN 60529)
Applicable standards: EN 60079-0: 2012+A11: 2013, EN 60079-1: 2007, EN 60079-28: 2007,
EN 60079-31: 2014, EN 60079-28: 2015
TDLS8000-E1 (IECEx)
Type of protection: Ex d [op is T6 Ga] IIC T5 Gc Ex tb IIIC T100ºC Db
Enclosure rating:
IP66 (In accordance with IEC 60529)
Applicable standards: IEC 60079-0: 2011, IEC 60079-1: 2007-04, IEC 60079-28: 2015, IEC
60079-31: 2013
Division 2, Zone 2: Nonincendive/Type n
TDLS8000-D2 (FM Approval for US)
Division system:
Type of protection: Nonincendive for Class I, Division 2, Groups A, B, C, D, T5
Dust-Ignitionproof for Class II/III, Division 1, Groups E, F, G, T5
Enclosure rating: Type 4X
Applicable standards: FM Class 3600: 2011, FM Class 3611: 2004, FM Class 3616:
2011, FM Class 3810: 2005
NEMA 250: 2003
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<2. Specifications>
2-6
Zone system:
Type of protection: Class I, Zone 2, AEx nA nC IIC T5
Zone 21, AEx tb IIIC T100ºC
Enclosure Rating: IP66
Applicable standards: ANSI/ISA-60079-0-2013, ANSI/ISA-60079-15-2012, ANSI/ISA60079-31-2015, ANSI/IEC 60529-2004 (R2011)
TDLS8000-C2 (FM Approval for Canada)
Type of protection: Ex nA nC IIC T5
Class II/III, Division 1, Groups E, F, G
Enclosure rating:
IP66, Type 4X
Applicable standards: CAN/CSA-C22.2 No.0-10 (R2015), CAN/CSA-C22.2 No.25-1966
(R2014), CAN/CSA-C22.2 No.94.1-07 (R2012), C22.2 No.94.2-07
(R2012), CAN/CSA-C22.2 No.60079-0:11, CAN/CSA-C22.2 No.6007915:12, CAN/CSA-C22.2 No.61010-1-12, CAN/CSA-C22.2 No.61010-2030-12, CAN/CSA-C22.2 No.60529-05 (R2010)
ANSI/ISA-12.27.01-2011
TDLS8000-S2 (ATEX)
Type of protection: II 3(1) G Ex nA nC [op is T6 Ga] IIC T5 Gc
II 2 D Ex tb IIIC T100 ºC Db
Enclosure rating:
IP66 (In accordance with EN 60529)
Applicable standards: EN 60079-0: 2012+A11: 2013, EN 60079-15: 2010, EN 60079-28: 2007,
EN 60079-28: 2015, EN 60079-31: 2014
TDLS8000-E2 (IECEx)
Type of protection: Ex nA nC [op is T6 Ga] IIC T5 Gc
Ex tb IIIC T100ºC Db
Enclosure rating:
IP66 (In accordance with IEC 60529)
Applicable standards: IEC 60079-0: 2011, IEC 60079-15: 2010, IEC 60079-28: 2015, IEC
60079-31: 2013
2.2
Others
2.2.1
YH8000 HMI Unit
The YH8000 is an HMI designed specifically for the TDLS8000. The YH8000 features an easyto-use touchscreen 7.5 inch color LCD which can be used to display maintenance information,
display alarm statuses and records, and set all parameters of the TDLS8000.
The YH8000 can be installed directly on the TDLS8000 or installed remotely.
An Ethernet connection is used to connect the YH8000 to up to four TDLS8000s simultaneously
via a hub.
l General Specifications
Display: Touchscreen 7.5 inch TFT color LCD panel, 640 x 480 (VGA)
Communication: Ethernet; RJ-45 connector
Communication speed; 100 Mbps
Case: Aluminum alloy
Paint color: Mint green (RAL 190 30 15 or equivalent)
Protection degree of enclosure:
IP65, Type 4X
Window: Polycarbonate
Weight: 4 kg
Mounting: Analyzer mount (Front, left-side, right-side) with tilt function, Pipe mount, or Panel
mount (Stainless steel)
Cable Entries: 1/2NPT or M20x1.5 mm, two holes
Installation conditions:
Ambient operating temperature;
-20 to 55ºC
Storage temperature:
-30 to 70ºC
Humidity: 10 to 90%RH at 40ºC (Non-condensing)
Power Supply: 24V DC +/-10%
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
2-7
<2. Specifications>
Power consumption: Max.12 W
Safety and EMC conforming standards:
Safety Conforming Standards:
CE EN61010-1
UL UL61010-1
CSA
CAN/CSA-C22.2 No.61010-1
GB
GB30439 Part 1
Installation Altitude: 2000 m or less
Installation category: I (Anticipated transient overvoltage 330 V)
Pollution degree:
2, Indoor/Outdoor use
EMC Conformity Standards:
CE
EN55011 Class A Group 1
EN61326-1 Class A Table 2 (For use in industrial location)
RCM
EN55011 Class A Group 1
KC
KN11 Class A Group 1, KN61000-6-2 (Korea Electromagnetic Conformity)
l Hazardous area classifications
Division 2, Zone2: Nonincendive/Type n
YH8000-D2 (FM Approval for US)
Division system
Type of protection: Nonincendive for Class I, Division 2, Groups A, B, C, D, T5
Enclosure rating: Type 4X
Applicable standards: FM Class 3600: 2011, FM Class 3611: 2004, FM Class 3810:
2005, NEMA 250: 2003
Zone system
Type of protection: Class I, Zone 2, AEx nA ic IIC T5
Enclosure rating: IP65
Applicable standards: ANSI/ISA-60079-0-2013, ANSI/ISA-60079-11-2014, ANSI/ISA60079-15-2012, ANSI/IEC 60529-2004 (R2011)
YH8000-C2 (FM Approval for Canada)
Type of protection: Ex nA nL IIC T5
Enclosure rating:
IP65, Type 4X
Applicable standards: CAN/CSA-C22.2 No. 0-10 (R2015), CAN/CSA-C22.2 No. 94.1-07
(R2012), CAN/CSA-C22.2 No. 94.2-07 (R2012), CAN/CSA-C22.2
No.60079-0:11, CAN/CSA-C22.2 No.60079-15:12, CAN/CSA-C22.2
No.61010-1-12, CAN/CSA No.60529-05 (R2010)
YH8000-S2 (ATEX)
Type of protection: II 3 G Ex nA ic IIC T5 Gc
Enclosure rating:
IP65 (In accordance with EN 60529)
Applicable standards: EN 60079-0: 2012+A11: 2013, EN 60079-11: 2012, EN 60079-15: 2010
YH8000-E2 (IECEx)
Type of protection: Ex nA ic IIC T5 Gc
Enclosure rating:
IP65 (In accordance with IEC 60529)
Applicable standards: IEC 60079-0: 2011, IEC 60079-11: 2011, IEC 60079-15: 2010
2.2.2
IF8000 Isolation Flanges
A process isolation flange protects the TDLS8000 from the process gas pressure and the heat,
dust, and corrosive elements of the process gas. A process isolation flange must be installed in
the following situations.
• When the process gas pressure exceeds 500 kPa
• When the process temperature is high and the temperature of the process window area
exceeds 55ºC even when process window purge is performed.
• When the process dust level is high and the adherence of dust or intrusion of corrosive
elements cannot be prevented even when process window purge is performed.
The IF8000 isolation flanges can be used for additional protection in in-situ or bypass
installations.
Note: Must use in conjunction with alignment flanges
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
2-8
<2. Specifications>
l General Specifications
Process connections: (See next table)
Heatresistance temperature: 200ºC max
Measured gas pressure:
Max. 1 MPa abs.
Wetted materials: Sapphire, 316 SS, Monel 400, Kalrez (O-ring)
Weight;
Process connection
Analyzer connection
ANSI Class 150-2-RF Flange
ANSI Class 300-2-RF Flange
ANSI Class 150-3-RF Flange ANSI Class 150-2-RF Flange
ANSI Class 300-3-RF Flange
ANSI Class 150-4-RF Flange
DIN PN16-DN50 Flange
DIN PN16-DN80 Flange
DIN PN16-DN50 Flange
JIS 10K-50-FF Flange
JIS 10K-80-FF Flange
Weight
316SS
Monel 400
5 kg/pc
6 kg/pc
7 kg/pc
7 kg/pc
8 kg/pc
9 kg/pc
11 kg/pc
12 kg/pc
12 kg/pc
14 kg/pc
7 kg/pc
7 kg/pc
10 kg/pc
11 kg/pc
7 kg/pc
7 kg/pc
9 kg/pc
10 kg/pc
Note: When using TDLS8000 as CE marking compliance product, the upper limit of the measurement gas pressure is 50kPa in gauge
pressure.
2.2.3
YC8000 Flow Cell
Used for extracting sample streams at any location.
Note: Must use in conjunction with alignment flanges (“-FC”)
l General Specifications
Gas temperature: 200ºC max
Gas pressure: Max. 1 MPa abs.
Wetted materials: Sapphire, 316 SS, Monel 400, Kalrez (O-ring)
Weight;
Material/Optical Path Length 1016 mm (40 inch)
Monel 400
15 kg
316SS
14 kg
1524 mm(60 inch)
18 kg
17 kg
Note: When using TDLS8000 as CE marking compliance product, the upper limit of the measurement gas pressure in YC8000 is
50kPa in gauge pressure.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
2.2.4
2-9
<2. Specifications>
Calibration Cell
Used for off-line calibrations and validations. Appropriate process windows are included on
calibration cell.
l General Specifications
Optical Path Length: 660 mm
Material: 316SS
Part No.
Description
K9772XA Calibration Cell with free-standing frame for O2
K9772XB Calibration Cell with free-standing frame for O2 LAO
K9772XC Calibration Cell with free-standing frame for ppm H2O in non-hydrocarbon
background
K9772XD Calibration Cell with free-standing frame for NH3
K9772XE Calibration Cell with free-standing frame for ppm H2O in hydrocarbon
background
K9772XF Calibration Cell with free-standing frame for ppm CO
K9772XG Calibration Cell with free-standing frame for ppm CO LAO
K9772XH Calibration Cell with free-standing frame for CO (%) + CO2 (%), CO2 (%)
Extend. Range
K9772XJ Calibration Cell with free-standing frame for HCl
K9772XL Calibration Cell with free-standing frame for CO(%), CO2 (%) High Range
K9772XM Calibration Cell with free-standing frame for H2S
Weight
Approx.
14 kg
Note: When using TDLS8000 as CE marking compliance product, the upper limit of gas pressure in calibration cell is 50kPa in gauge
pressure.
2.2.5
Unit Connection Cable
Use for interconnecting the Sensor Control Unit and the Laser Unit.
l General Specifications
Construction: Double-shielded (Overall shield and Individual shields) 4-pair cable
Part No.
K9775WA
K9775WB
K9775WC
K9775WD
K9775WE
K9775WF
K9775WG
Cable length
5m
10 m
20 m
30 m
40 m
50 m
60 m
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<2. Specifications>
2.3
2-10
Model and Codes
l TDLS8000 Tunable Diode Laser Spectrometer
Model
Suffix Code
Option Code
TDLS8000 ................................. ....................
....................
Type
-G1
....................
-G2
....................
-D2
....................
-C2
....................
-S2
....................
-E2
....................
-D1
....................
-C1
....................
-S1
....................
-E1
....................
Gas
-X1
....................
Parameter
-X2
....................
-C1
....................
-C2
....................
-C3
....................
-C4
....................
-C5
....................
-A1
....................
-A2
....................
-S1
....................
-D1
....................
-D5
....................
-H1
....................
-H3
....................
-H4
....................
-L1
....................
Optics Accessory -NN
....................
-LA
....................
-U2
....................
-U3
....................
-U4
....................
-D5
....................
-D8
....................
-J5
....................
-J8
....................
-FC
I/O Interface
-A1
....................
SI Unit
-J
....................
-N
....................
—
-N ....................
Option
/D
/RX
/RC
/SCT
*1:
*2:
*3:
*4:
*5:
*6:
*7:
*8:
*9:
Description
Tunable Diode Laser Spectrometer
General Purpose, cable entry/piping:NPT
General Purpose, cable entry:Metric thread, piping:Rc
FM (US) Class I Div 2, Zone2, cable entry/piping:NPT
FM (Canada) Class I Zone2, cable entry/piping:NPT
ATEX Type of protection “n”, cable entry:Metric thread, piping:Rc
IECEx Type of protection “n”, cable entry:Metric thread, piping:Rc
FM (US) Class I Div 1, Zone1, cable entry/piping:NPT
FM (Canada) Class I Zone1, cable entry/piping:NPT
ATEX Flameproof ”d”, cable entry:Metric thread, piping:Rc
IECEx Flameproof “d”, cable entry:Metric thread, piping:Rc
O2 < 600°C, 0-25%
O2 < 1500°C, 0-25% Combustion
CO (%) 0-20%/0-50% <500°C (*2)
CO ppm 0-200ppm/0-10,000ppm <500°C (*2) (*8)
CO ppm <1500°C Combustion (*8)
CO ppm <1500°C or CH4 0-5% Combustion (*8)
CO (%) + CO2(%) 0-30%/0-100% <150°C (*2)
NH3 up to 0-5,000ppm <450°C DeNOX
NH3 0-30ppm/0-5,000ppm + H2O 0-5%/0-50% <400°C (*2)
H2S 0-5%/0-100% <100°C (*2)
CO2 High Range 0-1%/0-5% <100°C (*2)
CO2 Extend. Range 0-30/0-50% <150°C (*2)
H2O ppm non-Hydrocarbon Background
H2O ppm Hydrocarbon Background
H2O 0-10%/0-100% <500°C (*2)
HCl 0-50ppm/0-5,000ppm <500°C (*2)
Without Alignment Flanges (*1)
Large Aperture Optics, ANSI CLASS150-4-RF (*2) (*3) (*7)
ANSI CLASS150-2-RF(Eq.) Alignment Flange, cable gland: NPT
ANSI CLASS150-3-RF(Eq.) Alignment Flange, cable gland: NPT
ANSI CLASS150-4-RF(Eq.) Alignment Flange, cable gland: NPT
DIN PN16-DN50-D(Eq.) Alignment Flange, cable gland: Rc
DIN PN16-DN80-D(Eq.) Alignment Flange, cable gland: Rc
JIS 10K-50-FF(Eq.) Alignment Flange, cable gland: Rc
JIS 10K-80-FF(Eq.) Alignment Flange, cable gland: Rc
Flow Cell Alignment Flange (*3)
Analog with HART+Modbus Ethernet
Only SI Unit
SI Unit or non SI Unit (*9)
Always -N
Diverging Beam without LAO (*4)
Reference Cell for O2 (*5)
Reference Cell for CO (*6)
Stainless Steel Tag Plate
When “-NN” is selected, Zone2/Div2/Type of protection “n”, FM (Canada) Zone1 is not available.
When “-LA” is selected, Zone1/Div1/Flameproof “d” is not available.
When FM (US) or FM (Canada) is specified, the connecting port for window purge is 1/4NPT.
When ATEX, IECEx or TIIS is specified, the connecting port for window purge is Rc1/4.
The Option “/D” can be selected when Large Aperture Optics “-LA” of the Optic Accessory is not specified and Oxygen or CO
(-C2, -C3, -C4) analyzer is selected.
The Option “/RX” can be used when Oxygen analyzer is selected. When both “-X2” of the Gas Parameter and “-LA” of the Optics
Accessory are selected, “/RX”must be specified.
The Option “/RC” can be used when CO analyzer is selected. When both “-C3” or “-C4”of the Gas Parameter is selected, “/RC”
must be specified.
For applications whose optical path length is 6 m or longer, select the “-LA”. A condensing lens unit (LAO unit) is added to the
SCU side.
When CO or CH4 ingredient coexist, please contact with YOKOGAWA.
An end user is available for case choice except Japan.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
2-11
<2. Specifications>
l YH8000 HMI Unit
Model
YH8000
Type
Language
—
Option
*1:
Suffix Code
...................
-G1
-G2
-D2
-C2
-S2
-E2
-E
-N
Option Code
....................
....................
....................
....................
....................
....................
....................
....................
....................
/M
/P
/W
/S
/C
/SCT
Description
HMI Unit
General Purpose, NPT thread for cable entry
General Purpose, Metric thread for cable entry
FM (US) Class I Div 2, Zone2, NPT thread for cable entry
FM (Canada) Class I Zone2, NPT thread for cable entry
ATEX Type of protection “n”, Metric thread for cable entry
IECEx Type of protection “n”, Metric thread for cable entry
English and 9 languages (*1)
Always -N
Mounting kit for TDLS8000
Pipe mount
Wall mount
Sun Shield
Local HMI connection cable: 3m
Stainless Steel Tag Plate
These languages are message languages on the display.
One analyzer has English and 9 languages.
All languages are as follows; English, German, French, Spanish, Portuguese, Russian, Hungarian, Korean, Chinese and
Japanese.
l IF8000 Isolation Flanges
Model
IF8000
Process
Connection
(*2)
Suffix Code
.............................................
-21
-23
-31
-33
-41
-50
-80
-J5
-J8
Analyzer
-21
Connection (*3)
-50
Material
-MN
-SS
Sapphire Window Type
-12
-13
-14
-15
-16
-17
-18
-20
—
-N
*1:
*2:
*3:
Option Code
....................
....................
....................
....................
....................
....................
....................
....................
....................
....................
....................
....................
....................
....................
....................
....................
....................
....................
....................
....................
....................
....................
....................
Description
Isolation Flange for TDLS8000 (2pcs/unit) (*1)
ANSI CLASS150-2-RF(Eq.)
ANSI CLASS300-2-RF(Eq.)
ANSI CLASS150-3-RF(Eq.)
ANSI CLASS300-3-RF(Eq.)
ANSI CLASS150-4-RF(Eq.)
DIN PN16-DN50-D(Eq.)
DIN PN16-DN80-D(Eq.)
JIS 10K-50-FF(Eq.)
JIS 10K-80-FF(Eq.)
ANSI CLASS150-2-RF(Eq.)
DIN PN16-DN50-D(Eq.)
Monel 400
316/316L SS
Coated for O2 (-X1, -X2)
Coated for ppmH2O non Hydrocarbon background (-H1)
Coated for ppmNH3 (-A1, -A2)
Coated for ppmH2O Hydrocarbon background (-H3)
Coated for ppmCO (-C2, -C3, -C4)
Coated for %CO or %CO2 (-C5, -D5)
Coated for HCl (-L1)
Coated for -C1, -D1, -H4, -S1
Always -N
IF8000 is delivered with two sets (for LU and SCU).
When ANSI flange of the Process Connection is selected, the “-21” of Analyzer Connection must be specified.
When DIN or JIS of the Process Connection is selected, the “-50” of Analyzer Connection must be specified.
The Analyzer Connection must be selected according to the flange size of TDLS8000.
IM 11Y01D01-01EN
3rd Edition: Jun. 15, 2016-01
<2. Specifications>
2-12
l YC8000 Flow Cell
Model
Suffix Code
YC8000
................................................................
Flow Cell Type -EN
Optical Path Length -40
-60
Material
-MN
-SS
Port Configuration
-3
Window Type
-XX
-H3
-HH
-NH
-CC
-C2
-HC
-MC
Inside Wall Treatment
-NN
-EP
—
-N
Option Code
....................
....................
....................
....................
....................
....................
....................
....................
....................
....................
....................
....................
....................
....................
....................
....................
....................
....................
Description
Flow Cell for TDLS8000
Enhanced
Forty Inches
Sixty Inches
Monel 400
316/316L SS
3 ports
Oxygen (-X1, -X2)
Moisture Hydrocarbon background (-H3)
Moisture non Hydrocarbon background (-H1)
NH3 (-A1, -A2)
ppmCO (-C2, -C3, -C4)
CO%+CO2% (-C5, -D5)
HCl (-L1)
-C1, -D1, -H4, -S1
No treatment (cleaned)
Electro-polish
Always -N
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
2.4
2-13
<2. Specifications>
External Dimensions
n TDLS8000 with Alignment Flange
l LU
Unit: mm
332
17
Alignment flange *
20
Measured
gas
198
Purge port (OUT) x2
1/4NPT or Rc1/4
200
Ø180
33
*:The alignment flange varies according to specifications.
Other flanges may be added.
SCU cable gland
3/4NPT or M25x1.5
237
33
Purge port (IN)
1/4NPT or Rc1/4
80
34
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
2-14
<2. Specifications>
l SCU
Unit: mm
332
Alignment
flange *
17
20
Measured
gas
198
62
Purge port (OUT) x2
1/4NPT or Rc1/4
196
LU cable gland
3/4NPT or M25x1.5
*:The alignment flange varies according to specifications.
Other flanges may be added.
Ø180
65
For I/O signal
3/4NPT or M25x1.5
For power supply
3/4NPT or M25x1.5
For YH8000
1/2NPT or M20x1.5
Purge port (IN)
1/4NPT or Rc1/4
237
33
80
34
l Maintenance space
Unit: mm
150
150
150
600
250
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
2-15
<2. Specifications>
l Alignment Flange
A (variable)
Purge port (IN) x2
1/4NPT or Rc1/4
Unit: mm
Q-h
t
ØC
Ø61.7
ØD
TDLS8000 side
Process side
Optics Accessory code
Hole QTY
(flange)
Q
-U2 ANSI CLASS150-2-RF(Eq.)
4
-U3 ANSI CLASS150-3-RF(Eq.)
4
-U4 ANSI CLASS150-4-RF(Eq.)
8
-D5 DIN PN16-DN50-D(Eq.)
4
-D8 DIN PN16-DN80-D(Eq.)
8
-J5 JIS 10K-50-FF(Eq.)
4
-J8 JIS 10K-80-FF(Eq.)
8
Hole
h
19
19
19
18
18
19
19
Hole P.C.D Thickness Outside dia. Distance
Purge port
C
t
D
A
120.7
19.5
150
87
1/4NPT
152.4
24.3
190
92
1/4NPT
190.5
23.9
228.6
92
1/4NPT
125
18
165
86
Rc1/4
160
20
200
88
Rc1/4
120
16
155
84
Rc1/4
150
18
185
86
Rc1/4
l LAO (Large Aperture Optics); Optics Accessory code “-LA”
This accessory is only for SCU side. For LU side, the Alignment flange ANSI CLASS150-4-RF
(Eq.) will be mounted.
When piping is Rc1/4, a conversion adapter will be attached on the Alignment flange of the LU
side.
23.9
Ø108.2
Purge port (IN) x2
1/4NPT or
Rc1/4 (with adapter)
Unit: mm
Ø229
Ø190.5
55
Process side
189 (Variable)
TDLS8000 side
8-Ø19 holes
Instalation for process side
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
2-16
<2. Specifications>
n YH8000 HMI Unit
Cable entry 2 (for Modbus)
1/2NPT or M20x1.5
261
65
204
Earth terminal
87.5
Unit: mm
75
75
4 - M6 depth 13
Cable entry 1 (for TDLS8000)
1/2NPT or M20x1.5
l Mounting kit for TDLS8000 (Option code: /M)
185
261
Unit: mm
440
(198)
Front mounting
302
337
(198)
(200)
Right mounting
Available for left mounting
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
2-17
<2. Specifications>
l Pipe mount (Option code: /P)
261
Unit: mm
108.3
138
123
54.3
46.3
Bracket
for pipe mount
102
204
65
110
2B pipe
Virtical or horizontal mounting
87.5
Sun Shield (Option code: /S)
Unit: mm
197
31
318
46.3
Stud for
sun shield
Bracket
for pipe mount
212.5
2B pipe
Virtical or horizontal mounting
178
l Wall mount (Option code: /W)
10
266
246
105.4
Bracket for wall mount
246±0.5
Unit: mm
10
226 206
206
±0.5
4 - Ø6.0 to 7.2 holes
or M6
Instalation for wall
4 - Ø7.2 holes
*:
The wall construction for mounting has to be designed for 4 times the weight of the YH8000.
Bracket for wall mount can be placed in lengthwise
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
2-18
<2. Specifications>
Sun Shield (Option code: /S)
318
10
196
246
206
10
Unit: mm
Bracket for wall mount
206±0.5
30
266 246
246
±0.5
4 - Ø6.0 to 7.2 holes
or M6
Instalation for wall
4 - Ø7.2 holes
When the sun shield is mounted, the bracket for wall have to be placed in widthwise.
n IF8000 Isolation Flanges
Unit: mm
Nut
5/8UNC, 3/4UNC
or M16
ØC
Q-h
Ø38
ØD
t
L
TDLS8000 side
Process side
Process Connection code
(flange)
-21
-23
-31
-33
-41
-50
-80
-J5
-J8
ANSI CLASS150-2-RF(Eq.)
ANSI CLASS300-2-RF(Eq.)
ANSI CLASS150-3-RF(Eq.)
ANSI CLASS300-3-RF(Eq.)
ANSI CLASS150-4-RF(Eq.)
DIN PN16-DN50-D(Eq.)
DIN PN16-DN80-D(Eq.)
JIS 10K-50-FF(Eq.)
JIS 10K-80-FF(Eq.)
Analyzer
Hole
Hole
Connection code QTY
h
(flange)
Q
4
19
ANSI
8
19
-21 CLASS150-24
19
8
22
RF(Eq.)
8
19
4
18
DIN PN168
18
-50
4
19
DN50-D(Eq.)
8
19
Purge port (IN) x2
1/4NPT or Rc1/4
Nut
5/8UNC
3/4UNC
5/8UNC
M16
Instalation for process side
Hole
Outside Bolt
Purge
Thickness
P.C.D
dia. length
port
t
C
D
L
120.7
39.6
150
127
127
39.6
165
137
152.4
39.6
190
137 1/4NPT
168.3
39.6
210
146
190.5
39.1
228.6
137
125
41.6
165
137
160
41.6
200
137
Rc1/4
120
40.6
165
139
150
40.6
185
139
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
2-19
<2. Specifications>
n YC8000 Flow Cell
TDLS8000 have to be assigned the dedicated Alignment flange (Optic Accessory: -FC).
When piping is Rc1/4, a conversion adopter will be attached on the Alignment flange.
1053 (Optical path length: -40) or 1561 (Optical path length: -60)
Unit: mm
Pipe size: 1-1/2 Sch80
Ø152.4
Sample inlet
(for 1/4” pipe)
86 (vaeiable) 35
Sample outlet
(for 1/4” pipe)
Gas temp. or
pressure measurement port
(for 1/4” pipe)
Ø152.4
Flow cell aligenment
flange
(TDLS8000
Optics accessory
code “-FC”)
n Calibration Cell
Part number: K9772XA, K9772XB, K9772XC, K9772XD, K9772XE, K9772XE, K9772XF,
K9772XG, K9772XH, K9772XJ, K9772XL, K9772XM
Unit: mm
Approx. 690
Sample port x2 (IN, OUT) 1/4NPT
Ø152.4
Pipe size: 1-1/2 Sch80
Approx.
410
Approx. 820
210
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
Blank Page
3.
3-1
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
Installation, Wiring, Optical Axis
Adjustment, and Piping
This chapter describes installation, wiring, optical axis adjustment, and purge gas piping in the
order they need be performed.
If you are going to use the YH8000, install it after you complete the procedures in this chapter.
3.1 Installation
The TDLS8000 uses a laser beam. After installation, the optical axis needs to be adjusted. Install
it in a location with sufficient working space.
CAUTION
During installation, be careful not to drop the product, damage the display, and so on.
Refer to “1.1 System configuration” and “2.3 Model and Codes” to install the product. Reserve
space for maintenance so that you can adjust the optical axis.
NOTE
Use a process isolation valve with an opening whose diameter is at least 38 mm so that you can
sufficiently adjust the optical axis of the laser beam after installation.
Flange alignment and installation are important. A proper flange installation ensures accurate
optical axis adjustment of the laser beam.
l Installation conditions
Install the product in a location that meets the conditions indicated in “2.1 TDLS8000 Tunable
Diode Laser Spectrometer Specifications”.
Note the following points.
• If an abnormal temperature occurs inside the case due to the influence of gas temperature,
attach a reducer or the like to separate the laser unit (LU) and sensor control unit (SCU)
from the flue.
• Process window purge protects the TDLS8000 from the heat, dust, and corrosive elements
of the process gas. Be sure to run the process window purge gas during processing. The
process window purge gas flow rate varies depending on the process gas conditions.
Temperature: Set the purge gas flow rate so that the temperature of the process
window area and in the inside of the alignment flange does not exceed 55 °C.
Dust:
Set the purge gas flow rate so that the transmission can be maintained. If the
transmission decreases over time, the purge flow rate must be increased.
Corrosion: If the process includes corrosive elements, sufficient purge flow rate is
necessary. If the sealant of the TDLS8000’s process window corrodes,
the inside of the TDLS8000 will also corrode, causing the TDLS8000 to
malfunction. Set the purge gas flow rate appropriately to keep corrosive gas
from entering the process window area or inside the alignment flange.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3.1.1
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
3-2
Measurement Point Selection
Take the following process conditions into consideration when selecting the measurement point.
l Process gas flow rate conditions
Set the measurement point to a location where the concentration distribution of the streamline
flow is uniform.
In the case of a duct or flue with a circular cross section, a typical measurement point is where
the distance from the end of a curved process area is at least three times the diameter (D) of the
duct or flue and where there is nothing that would interfere with measurements.
In the case of a duct or flue with a rectangular cross section, the equivalent diameter (D) can be
determined from the following equation.
Diameter (D) = 4 × duct cross sectional area/duct circumference
If such point is not available or if setting a measurement point at such point is not possible, the
measurement point is a location two-thirds of the length away from the duct inlet end or one-third
from the outlet end.
Once the measurement point is determined, double-check that it is at the appropriate location.
l Process gas temperature
Install the TDLS8000 in a location with minimal process gas temperature fluctuations.
If the gas temperature fluctuation where the TDLS8000 is installed exceeds ±10 °C, connect
an external thermometer to the TDLS8000 temperature input terminal and enter the actual
measured gas temperature to obtain correct measurements (for details, see “6.1.3
Process
Temperature”).
Check that a thermometer suitable for the maximum process gas temperature is being used.
In general, the lower the gas temperature, the better the measurement.
l Process gas pressure
Install the TDLS8000 in a location with minimal pressure fluctuations.
If the gas pressure fluctuation where the TDLS8000 is installed exceeds ±5 kPa, enter
the pressure signal from a separately applied process pressure meter to obtain correct
measurements (for details, see “6.1.2 Process Pressure”).
Check that a pressure meter suitable for the maximum process gas pressure is being used.
Check that the process window interfacing the process gas is suitable for the maximum preset
gas pressure.
In general, the lower the gas pressure, the better the measurement.
l Process dust/particulate concentration
Install the TDLS8000 in a location with minimal dust concentration.
The transmission of the measurement laser beam will decrease due to dust or particles.
Measurement will not be affected as long as the transmission loss is within limits, but if it exceeds
limits, an alarm will occur. The dust load also depends on the process optical path length. For
details, contact your nearest Yokogawa representative.
In addition, install the TDLS8000 in the location such as not causing breakage of the process
window and LAO unit lens by foreign object might damage the window and lens in process gas.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3.1.2
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
3-3
Constructing Process Flanges
Please prepare your own process flanges.
n Process Flange Reinforcing Plate
If the duct or the wall of flue that the process flanges will be attached to is thin or may bend, weld
large reinforcing plates around the area where the flanges will be attached. Figure 3.1 shows an
example of welding reinforcing plates. Provide a secure platform to install the TDLS8000 on at
your discretion.
Approx. 10 mm thick steel plate (welded)
Reinforcing plate
(welded)
LU or SCU
Side view
Figure 3.1
Wall containing the measured gas
Front view
Reinforcing plate for LU and SCU process flanges
CAUTION
Duct or flue with thin and easily bent walls must be reinforced to maintain laser beam alignment
at all times. Reinforce the process flanges to maintain alignment.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3-4
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
n Alignment Angle Tolerance
The LU and SCU have alignment mechanisms that can be used to manually adjust the laser
beam direction within vertical and horizontal planes. Make sure that the process flanges are
within the tolerance angles indicated in Figure 3.2.
90º ±2º
90° ±2°
Measured gas (duct, wall of flue, etc.)
Attachment angle tolerance (top view)
45
°
±2°
Bolt hole position (front view)
Attachment angle tolerance (side view)
Attach the flange at the flange attachment
bolt hole position shown in the figure
above so that the LU and SCU of the
TDLS8000 can be attached in the
correct orientation.
Attach the flanges so that the total angle
deviation of the flange nozzles on the LU
end and SCU end is within ±2° in either
direction.
Figure 3.2
Alignment angle tolerance of LU and SCU flanges
n Process Flange Opening
Attach LU and SCU flanges and insertion tubes (when used) so that the TDLS8000 can safely
and easily be installed, operated, and removed.
NOTE
If the laser unit and sensor control unit are far apart, a larger opening may be required. For
details, contact your nearest Yokogawa representative.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3-5
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
45 mm
Minimum allowable opening
Top view
Measured gas (duct, wall of flue, etc.)
45mm
Minimum allowable opening
Side view
Figure 3.3
To secure the minimum allowable
opening, minimize the horizontal
axis deviation of the flange nozzles.
Openings for LU and SCU flanges
n Selecting the Appropriate Process Flanges
(1) Process flanges with optical path length less than 6 m
Use process flanges that conform to any of the flange standards listed in chapter 2.
(2) Process flanges with optical path length greater than or equal to 6 m
The optical axis tends to deviate with longer process optical path lengths. As such, an LAO unit
will be required as a process interface.
NOTE
When an LAO unit is used with a flange with process optical path length greater than or equal to
6 m, only a process flange conforming to ASMEB16.5 4”-Class150(RF) can be used.
3.1.3
Attaching the TDLS8000 to the Process Flange
The installation involves fixing an alignment flange or the like to a process flange first and then
attaching the TDLS8000.
If the process conditions include high temperature, high pressure, or corrosive gas, insert a
process isolation flange (IF8000) between the process flange and alignment flange.
If there is a lot of dust, use an insertion tube. The insertion tube is an interface for preventing dust
from adhering to the TDLS8000 process window. The length is determined by the process flange
length, the dust level in the process, the process gas flow velocity, and so on. If you require an
insertion tube, please consult with Yokogawa.
For details of each component, see “1.1 System configuration” and “2.4 External Dimensions”.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3-6
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
n Attaching Alignment Flanges
To attach an alignment flange to a process flange, follow the procedure below.
(1) Insert a gasket between the process flange and alignment flange.
(2) Pass the bolts through the alignment flange mounting holes, and fasten with nuts. Make
sure that the nuts are securely fastened so that they do not fall off.
CAUTION
• The alignment flange must be installed in the proper orientation. Install it so that the arrows
of “↑UP↑” are facing up.
• The process window area is where the laser beam passes through. Be careful not to
damage or stain the window during installation.
Alignment flange
Nut
Figure 3.4
Gasket
Process flange
Mounting hole Mounting hole Bolt
(4 or 8 locations) (4 or 8 locations)
Installing an alignment flange
n Attaching Process Isolation Flanges (IF8000)
To attach a process isolation flange (IF8000) and an alignment flange to a process flange, follow
the procedure below.
(1) Insert a gasket between the process flange and process isolation flange.
(2) Secure the process isolation flange to the process flange with four or eight bolts. The
process isolation flange must be installed in the proper orientation. Install it so that the
arrows of “↑UP↑” are facing up.
(3) Insert a gasket between the TDLS8000 side of the process isolation flange and alignment
flange.
(4) Pass the alignment flange through the bolts welded to the process isolation flange, and
fasten with nuts. Make sure that the nuts are securely fastened.
CAUTION
• The alignment flange must be installed in the proper orientation. Install it so that the arrows
of “↑UP↑” are facing up.
• The process window area is where the laser beam passes through. Be careful not to
damage or stain the window during installation.
• If there are welded bolts (process isolation flange), be careful not to crush or damage the
threads during installation.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3-7
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
Alignment flange
Nut
Figure 3.5
Gasket
Process isolation flange
Mounting hole
(4 or 8 locations)
Gasket
Process flange
Mounting hole
(4 or 8 locations)
Nut
Installing a process isolation flange and alignment flange
n Attaching Insertion Tubes
The shape of the insertion tube used when there is a lot of dust may vary from that in the figure
depending on the measurement conditions.
CAUTION
• The alignment flange must be installed in the proper orientation. Install it so that the arrows
of “↑UP↑” are facing up.
• The process window area is where the laser beam passes through. If there is a process
window area (alignment flange, process isolation flange), be careful not to damage or stain
it during installation.
• If there are welded bolts (process isolation flange, insertion tube), be careful not to crush or
damage the threads during installation.
l When only an alignment flange is to be used (IF8000 is not used)
(1) Insert a gasket between the process flange and the process side of the insertion tube.
(2) Secure the insertion tube to the process flange with four or eight bolts.
(3) Insert a gasket between the TDLS8000 side of the insertion tube and alignment flange.
(4) Pass the alignment flange through the bolts welded to the insertion tube, and fasten with
nuts. Make sure that the nuts are securely fastened.
Alignment flange
Nut
Figure 3.6
Gasket
Mounting hole
(4 or 8 locations)
Insertion tube
Gasket
Process flange
Mounting hole
(4 or 8 locations)
Nut
Installing an insertion tube and alignment flange
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3-8
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
l When a process isolation flange (IF8000) and alignment flange are to be
installed
(1) Insert a gasket between the process flange and the process side of the insertion tube.
(2) Insert a gasket between the insertion tube and process isolation flange.
(3) Insert a gasket between the TDLS8000 side of the process isolation flange and alignment
flange.
(4) Pass the bolts welded to the process isolation flange through the alignment flange mounting
holes and process flange mounting holes, and fasten with nuts. Make sure that the nuts are
securely fastened so that they do not fall off.
Alignment flange
Nut
Figure 3.7
Gasket
Mounting hole
(4 locations)
Process isolation flange
Gasket
Insertion tube
Gasket
Process flange
Mounting hole
(4 or 8 locations)
Nut
Installing an insertion tube, process isolation flange, and alignment flange
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3-9
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
n TDLS8000 Installation
CAUTION
The process window is where the laser beam passes through. Be careful not to damage or stain
the window during installation.
After mounting an alignment flange and the like to the process flange, install the TDLS8000
according to the following procedure.
(1) First fasten three M6 screws in the TDLS8000 mounting screw holes on the flange surface
of the alignment flange. Leave about 8 mm of gap from the flange surface. Do not fasten the
screw in the upper right hole as viewed from the front. The upper right screw is attached on
the TDLS8000 side.
(2) Insert the alignment flange screws that you fastened in (1) in the holes on the mounting
surface (quick connectors) of the laser unit (or sensor control unit), and then rotate the unit
clockwise.
(3) Temporarily fasten the upper right screw, and then tighten all screws evenly.
CAUTION
Carefully mount the laser unit (or sensor control unit) to prevent the O-ring from being damaged
or dropping out during assembly.
CAUTION
Anti-seizing grease is applied to the screw areas. Be careful to keep dust or the like from
adhering. If dust or the like adheres to the areas, remove the dust or the like, and then reapply
the anti-seizing grease.
Hole with a hexagonal hole
Quick connector
Turn clockwise
Process flange
Alignment flange
Figure 3.8
SCU or LU installation
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
3-10
3.2 Wiring
When the installation is complete, wire the TDLS8000 and external devices.
YH8000 wiring is explained in chapter 4.
n Wiring Precautions
To open the SCU and LU covers, turn the lock screw counterclockwise with the supplied hex
wrench to loosen the screw.
After closing the SCU and LU covers, turn the lock screw clockwise to tighten.
CAUTION
• Turning the cover without loosening the lock screw can damage the case or cover. Note that
the lock screw is in a loosened state when the TDLS8000 is shipped from the factory.
• If sand or foreign substance adheres to the screw area of the cover or case, wipe it off to
prevent it from damaging the screw threads and prevent it from entering the inside of the
device.
Lock screw
Cover
Hex wrench
Figure 3.9
CAUTION
Never turn on the power to the TDLS8000 or the devices connected to the TDLS8000 until all
wiring is complete.
l Wiring procedure
Construct signal cables and power supply cables according to the following conditions.
(1) Be sure to connect the shield to the functional ground terminal for the shielded wire inside
the TDLS8000.
(2) Strip the necessary minimum length of outer most covering of signal cables and power
supply cables.
(3) When using conduit tubes, do not run power cables in the same conduit as signal cables.
Doing so can cause noise interference on signals. Ground metal conduits.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3-11
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
(4) Attach the supplied blind plugs to unused cable glands.
(5) For the cables you need to use, see “n Types of Wiring and Cabling”.
(6) When you complete all wiring, close the terminal cover, and fasten with the lock screw.
n Cable Entries
Symbols are inscribed near the cable entries for identifying the thread specifications.
M
For ANSI 1/2NPT, 3/4NPT(F): A
For ISO M20x1.5, M25x1.5: M
Figure 3.10
Attach conduits and cable glands with the appropriate thread size to the TDLS8000 cable entries.
Cable entry 1
(3/4NPT
or M25)
Cable entry
(3/4NPT or M25)
Cable entry 2 Cable entry 3
(3/4NPT or M25)
Sensor control unit (SCU)
Laser unit (LU)
Figure 3.11
Cable entry 4
(1/2NPT
or M20)
Cable entries
Sensor control unit (SCU) cable entries
Cable entry 1:
Cable entry for inter-unit cables
Cable entry 2:
Cable entry for I/O signal cables
Cable entry 3:
Cable entry for power cables
Cable entry 4:
Cable entry for a cable connecting to the YH8000 or Ethernet cable
Laser unit (LU) cable entry
Cable entry:
Cable entry for inter-unit cables
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3-12
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
n TDLS8000 Wiring Terminals
CAUTION
Be careful not to connect the power supply wires to the incorrect locations or reverse the polarity.
In particular, incorrectly connecting the power supply terminals (POWER, VO, VO[HMI]) or
solenoid valve control output terminals (SV-1, 2) can damage the TDLS8000 or the devices
connected to the TDLS8000.
Laser Unit (LU)
Sensor Control Unit (SCU)
Ethernet port for
YH8000 or DCS
Terminal B
Terminal A
Terminal C
Earth for shield wire
Earth for shield wire
Earth terminal
Magnified Terminal B
Magnified Terminal A
LC
+
MS-1
-
+
-
MS-2
+
-
VO
+
AO-1
-
+
-
AO-2
+
-
Earth terminal
AI-1
+
AI-2
-
+
-
DI-1
+
-
DI-2
+
4-20mA
Isolated
Digital Input
Input for
4-20mA
Pressure
Output
With Isolated transmitter 4-20mA
HART 4-20mA
Input for
Output
Temperature
transmitter
Magnified Terminal C
Connect to shield wire terminal
(Both side of cable)
LC
+
MS-1
-
+
-
MS-2
+
-
VO
+
-
SV-1
+
-
DO
Digital
Output for
FAULT
Digital
Output for
programmable
DO
SV-2 POWER
+
-
FAULT
- NC COM NO NC COM
+
-
VO (HMI)
+ +
-
-
Solenoid
Power
Valve
Supply
Control for
24V DC
24V DC Output for
Auto Cal Solenoid
YH8000 Power
Valve
Control for
Auto Cal
Figure 3.12
Table 3.1
Terminal
block
A
TDLS8000 wiring
Terminals and functions (laser unit: LU)
Terminal
Function
name
NC
Not used
LC+
Inter-unit communication, control signal, power supply
LC Connect these terminals to the terminals with the same names on
MS-1+
terminal block C in the SCU.
MS-1MS-2+
MS-2VO+
VO-
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3-13
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
CAUTION
Do not wire the NC terminal.
Table 3.2
Terminal
block
B
C
Terminals and functions (sensor control unit: SCU)
Terminal
name
AO-1+
AO-1AO-2+
AO-2AI-1+
AI-1AI-2+
AI-2DI-1+
DI-1DI-2+
DI-2DO NC
DO COM
DO NO
FAULT NC
FAULT
COM
LC+
LCMS-1+
MS-1MS-2+
MS-2VO+
VOSV-1+
SV-1SV-2+
SV-2POWER+
POWERVO[HMI]+
VO[HMI]-
Function
Analog output 1 (4-20 mA)/HART communication port
Analog output 2 (4-20 mA)
Analog pressure signal input (4-20 mA). Connect to a pressure transmitter.
Analog temperature signal input (4-20 mA). Connect to a temperature transmitter.
Digital input 1 Voltage-free digital input terminal. Open: 100 kΩ or more, Closed:
200 Ω or less (including wiring resistance)
Digital input 2 Voltage-free digital input terminal. Open: 100 kΩ or more, Closed:
200 Ω or less (including wiring resistance)
Programmable digital output
Between NC and COM: Closed when the specified operating condition is met
Between NO and COM: Open when the specified operating condition is met
FAULT signal digital output
Closed when the device is operating normally; open when a fault occurs or when
the power is off
Inter-unit communication, control signal, power supply
Connect these terminals to the terminals with the same names on terminal block A
in the LU.
Solenoid valve control output 1. Output rating: 24 V DC, 0.5 A max.
Solenoid valve control output 2. Output rating: 24 V DC, 0.5 A max.
Power supply. 24 V DC ± 10%
Power supply terminal for the YH8000. 24 V DC
n Types of Wiring and Cabling
CAUTION
Use cables with a durable temperature of at least 70 °C.
CAUTION
Use cables that are appropriate for the environment that the product is installed in. Use cables
with an outer diameter that matches the cable gland that you are using.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
Nominal cross
sectional area,
conditions
1, LU
Inter-unit Cables of other selling
cable
for the TDLS8000
K9775WA to K9775WG
(select according to
cable length)
AWG18
4 pairs
Covering outer diameter
approx. 12 mm
3
Power
AWG18 to AWG12
supply
Two-core or threecore (when using the
functional ground
terminal inside the
device)
2
I/O cable Multi-core cable
Up to 21 cores
A terminal box or the
like is required when
branching the signals
externally.
4
YH8000 Special cable (YH8000
connection option)
cable
AWG24
4 pairs
Covering outer diameter
approx. 8.4 mm
Ethernet CAT.5e
cable
AWG24
4 pairs
100 m max.
Functional Functional AWG16 or more
grounding
ground
terminal
(external)
Cable
entry
Cable
type
Shield
Terminal
Required
Wire:
(individual
Shield: M4 screw
shields for each crimp-on terminal
pair and overall
shield)
3-14
Withstand
voltage, flame See
resistance
500 V or more 3.2.1
FT-4
Required
Wire:
Shield: M4 screw
crimp-on terminal
500 V or more 3.2.2
VW-1 or more
Required
Wire:
Shield: M4 screw
crimp-on terminal
500 V or more 3.2.3
VW-1 or more to
3.2.7
Required
Wire:
Shield: M4 screw
crimp-on terminal
500 V or more 4
FT-4
Required
Wire:
Shield: M4 screw
crimp-on terminal
VW-1 or more 3.2.8
Not required
M5 screw crimp-on
terminal
3.2.2
Use cables with outer diameters that match the conduits or cable glands that you are using.
3.2.1
Connecting between the Sensor Control Unit (SCU)
and Laser Unit (LU)
CAUTION
Be careful not to connect the power supply wires to the incorrect locations or reverse the polarity.
Incorrectly wiring the power supply terminals (VO) and signal terminals may burn out the
TDLS8000 circuitry.
To connect between the sensor control unit (SCU) and laser unit (LU), use the inter-unit cable of
other selling (K9775WA to K9775WG, select according to the required cable length). Connect the
inter-unit cable as shown in Figure 3.13.
For the terminating the inter-unit cable, see “Appendix 1 Constructing Unit Connection Cables”.
(1) Connect pairs 1 to 4 of the inter-unit cable to terminal A of the laser unit.
(2) Connect the shielded wire to the ground terminal inside the laser unit.
(3) Attach the supplied ferrite clamp to the cable as shown in Figure 3.14.
(4) Connect pairs 1 to 4 of the inter-unit cable to terminal C of the sensor control unit. Connect
the positive terminal of the LU side to the positive terminal of the SCU side and the negative
to the negative.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3-15
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
(5) Like the laser unit, connect the shielded wire to the ground terminal inside the sensor control
unit.
(6) Within the SCU and LU, attach the supplied ferrite clamps to the cable as shown in Figure
3.14. Be sure to attach them on both sides (SCU and LU).
Laser unit
Terminal A
LC MS-1 MS-2 VO
Sensor control unit
Terminal C
LC MS-1 MS-2 VO SV-1 SV-2 POW VO(HMI)
Internal
ground
terminal
Internal
ground
terminal
Figure 3.13
Connecting between the Sensor Control Unit (SCU) and Laser Unit (LU)
To the ground terminal
Ferrite clamp
Shielded wire
To the ground
terminal
Cable entry
Figure 3.14
3.2.2
Connecting the Power Cable and Grounding
CAUTION
Be careful not to connect the power supply wires to the incorrect locations or reverse the polarity.
Incorrectly connecting the power supply can cause the TDLS8000 to malfunction.
Use two-core or three-core shielded cable to wire the power supply.
For ground wiring, use the internal ground terminal or external ground terminal. If you want to use
the internal ground terminal, use a three-core power cable.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
3-16
Sensor control unit
Terminal C
LC MS-1 MS-2 VO SV-1 SV-2 POW VO(HMI)
External ground terminal
Internal ground terminal
24 V DC power supply
3.2.3
Connecting to Temperature and Pressure
Transmitters
This section explains the wiring for receiving current signals (4 to 20 mA DC) from a temperature
and pressure transmitters. Connect AI-1 to a pressure transmitter and AI-2 to a temperature
transmitter.
For analog input settings, see “6.3 Analog Input Settings”.
n Connection Preparation
To supply power to the transmitters from the TDLS8000, set the switch inside the sensor control
unit to Active AI. To supply power externally, set to Passive AI. If you want to connect to a 4-wire
system pressure meter or thermometer, set to Passive AI.
The factory default setting is Passive AI.
AI-1 (pressure signal)
AI-2 (temperature signal)
Applicable
switch
SW1
SW2
Switch state
External
Power supply from
power supply
the TDLS8000
Passive AI
Active AI
Passive AI
Active AI
Sensor control unit
Active AI
Passive AI
SW1
SW2
Figure 3.15
SW1 and SW2 settings
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3-17
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
CAUTION
Check that the TDLS8000 is turned off before setting SW1 or SW2. Otherwise, the TDLS8000
may be damaged.
n Connecting a Pressure Meter and Thermometer
Connect the analog output terminals of the transmitters as follows. The terminal polarity is the
same for Passive AI and Active AI.
Sensor control unit
Terminal B
AO-1 AO-2 AI-1 AI-2 DI-1 DI-2
+ - + -
Internal
ground terminal
DO FAULT
Temperature transmitter
+
Pressure transmitter
+
-
n When Connecting an External Power Supply Such as a Distributor
If you need to connect an external power supply such as a distributor to a 2-wire system
transmitter, connect it as follows. Set the switch to Passive AI.
Sensor control unit
Terminal B
AO-1 AO-2 AI-1 AI-2 DI-1 DI-2
Internal
ground terminal
DO FAULT
+
External power supply
24V DC
Pressure transmitter
+
-
l Wiring procedure
• For the cable type to use, see “n Types of Wiring and Cabling”.
• Be sure to ground the cable shield on the TDLS8000 side and on the other side.
• When supplying power to the transmitters from the TDLS8000, take into account the drop in
the transmitter supply voltage due to wiring resistance and the like.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3-18
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
CAUTION
Do not apply current exceeding the allowable value to AI. Doing so can cause a malfunction.
3.2.4
Wiring Analog Outputs (AO)
This section explains the wiring for transmitting concentration, transmission, and other analog
outputs to a recorder or other device. Only AO-1 supports HART communication.
For analog output settings, see “6.4
Analog Output Settings”.
Sensor control unit
Terminal B
AO-1 AO-2 AI-1 AI-2 DI-1 DI-2
+ - + -
Internal
ground
terminal
DO FAULT
Recorder 2
+
Recorder 1
+
-
l Wiring procedure
•
•
•
•
For the cable type to use, see “n Types of Wiring and Cabling”.
Be sure to ground the cable shield on the TDLS8000 side and on the other side.
For each output, keep the load resistance including the wiring resistance 550 Ω or less.
During HART communication, keep the load resistance including the wiring resistance
within the allowable load resistance range specified by the HART communication
specifications, which is 250 to 550 Ω. (AO-1 only)
CAUTION
Be careful not to reverse the polarity when wiring. Doing so can cause a malfunction.
3.2.5
Wiring Digital Outputs
The following digital outputs are available. Both contacts are voltage-free dry outputs
(mechanical relay digital outputs). The contact rating is 24 V DC 1 A for both contacts.
For digital output settings, see “6.5
Digital Output Settings”.
l DO digital output (DO)
A user-defined function can be assigned to this contact through configuration. It is a C-contact
(transfer contact) consisting of three terminals: COM, NC, and NO. It is always de-energized and
cannot be changed. The NC and NO markings on the terminals indicate the de-energized state.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
3-19
Contact state
Power off
Output on
Output off
State between the NO State between the NC
and COM terminals
and COM terminals
Open
Closed
Closed
Open
Open
Closed
l FAULT digital output (FAULT)
This contact transmits a signal when a fault occurs. It is an A-contact (make contact) consisting of
two terminals: COM and NC. It is always energized and cannot be changed. The NC marking on
the terminal indicates the energized state.
Contact state
Power off
Output on
Output off
State between the NC
and COM terminals
Open
Open
Closed
Sensor control unit
Terminal B
AO-1 AO-2 AI-1 AI-2 DI-1 DI-2
Internal ground terminal
DO FAULT
Annunciator or the like
l Wiring procedure
• For the cable type to use, see “n Types of Wiring and Cabling”.
• Be sure to ground the cable shield on the TDLS8000 side and on the other side.
• The contact rating is 24 V DC 1 A. Connect a load (e.g., indicator lamp, annunciator) that will
not cause these values to be exceeded.
• For the DO digital output, select whether to wire NC or NO depending on your application.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3.2.6
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
3-20
Wiring Digital Inputs
The TDLS8000 executes specified functions when it receives contact signals. There are two
inputs. Apply voltage-free contact signals. The digital input terminal outputs 5 V DC.
For digital input settings, see “6.7 Digital Input Settings”.
Sensor control unit
Terminal B
AO-1 AO-2 AI-1 AI-2 DI-1 DI-2
+ - + -
Internal ground terminal
DO FAULT
SW1
SW2
l Wiring procedure
• For the cable type to use, see “n Types of Wiring and Cabling”.
• Be sure to ground the cable shield on the TDLS8000 side and on the other side.
• The open and closed levels of the digital inputs are identified by the resistance seen from
the TDLS8000 side. Wire the digital inputs to meet the following conditions. Note that wiring
resistance is included.
If the DI terminals are shorted, about 2 mA of current will flow.
Resistance
3.2.7
Closed
200Ω or less
Open
100 kΩ or more
Wiring Solenoid Valve Control Outputs
These outputs control the solenoid valves that are used during calibration and the like. There are
two outputs. Each can supply 24 V DC 500 mA max.
Sensor control unit
Terminal C
LC MS-1 MS-2 VO SV-1 SV-2 POW VO(HMI)
Internal ground terminal
Solenoid valve 1
Connect to the TDLS8000
shielded wire terminal.
Solenoid valve 2
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
3-21
l Wiring procedure
• For the cable type to use, see “n Types of Wiring and Cabling”.
• Be sure to ground the cable shield on the TDLS8000 side and on the other side.
• The output rating is +24 V DC 500 mA max. Check that the solenoid valves that you want to
use do not exceed these values before connecting them.
CAUTION
Do not short the SV terminals when the solenoid valve control output is on. Doing so will cause
the internal protection fuse to melt, preventing output. If this happens, the component needs to be
replaced.
3.2.8
Connecting an Ethernet Cable
NOTE
Reception of numerous invalid packets may affect the TDLS8000 functionality.
When connecting the TDLS8000 to a network, we recommended that you manage the network
appropriately.
If you want to connect the TDLS8000 to an YH8000 (HMI unit) through an Ethernet hub or to an
external device through Modbus/TCP communication, you will need to use an Ethernet cable.
The Ethernet cable connector must be crimped during the TDLS8000 installation.
NOTE
• Before crimping the Ethernet connector, pass the cable through cable gland. After crimping
the Ethernet connector, the connector cannot be passed through the cable glands.
• Be careful of the cable gland orientation. The end with the screw section of the cable gland
is the connector end.
• Use an eight-core CAT 5e shielded cable for the Ethernet cable.
Use a braided wire type shield. If the shield is a metallic foil type, the shield may not be
properly grounded.
Use a cable with straight wiring.
• Use a cable gland with a cable diameter specification that matches the outer diameter of the
Ethernet cable.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3-22
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
• The required components and tools are shown below.
RJ45 modular plug
RJ45 modular plug crimp tool
LAN cable tester
Wire cutter
Wire stripper
Round crimp-on terminal
For M4 screw
(for shielded wires)
Use the appropriate crimp-on terminals for the Ethernet cable that you are
using.
Crimp-on terminal example:
For M4 screws, nominal cross sectional area of wire 2 mm2
FV2-4 by J.S.T. Mfg. Co.,Ltd. or 170782-1 by TE Connectivity, or equivalent
Crimp tool for round crimp-on
terminals
Heat shrink tube
Used to cover the shielded parts of the Ethernet cable.
Use the appropriate heat shrink tube for the Ethernet cable that you are using.
Heat shrink tube example:
For shielded wires: inner diameter 4 mm, length about 140 mm
For external cable covering: inner diameter 10 mm, length about 30 mm
Heating gun
For shrinking heat shrink tubes
l Processing an Ethernet cable
(1) Pass the Ethernet cable through the cable gland.
(2) Remove about 150 mm of covering from the shielded Ethernet cable.
Be careful not to cut off the shield.
Outer covering
Cable gland
About 150 mm
(3) Bundle the shield in a linear form, cover it with a heat shrink tube, and apply heat treatment
to shrink the tube.
Then, attach a round crimp-on terminal to the end of the shield.
Cover with a heat shrink tube the section of the cable where the covering was removed, and
apply heat treatment to shrink the tube.
Bundle the shield in a linear form.
Outer covering
Cover with a heat shrink tube, and apply heat
treatment to shrink the tube.
Then, attach a round crimp-on terminal to the end.
Cover with a heat shrink tube, and apply heat treatment to shrink the tube.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
3-23
(4) Crimp an RJ45 modular plug onto the end of the Ethernet cable.
Pin No.
1
2
3
4
5
6
7
8
Wire color
White-orange
Orange
White-green
Blue
White-blue
Green
White-brown
Brown
Insert the wires in the correct arrangement into the RJ45 modular plug, check that the wires
are firmly inserted, and then crimp the plug with a RJ45 modular plug crimp tool.
Finally, check that the plug has been crimped on properly by testing the connection with a
LAN cable tester.
(5) Insert the RJ45 connector into the TDLS8000 Ethernet port and the round crimp-on terminal
at the end of the shielded wire to the functional ground terminal (M4 screw) inside the SCU.
3.3
Optical Axis Adjustment
When wiring is complete, turn on the power, and adjust the optical axis.
CAUTION
The TDLS8000 is a Class 1 laser product. As such, the laser level of the product is safe to the
eyes, but do not intentionally look at the laser light source. The TDLS8000 laser unit emits laser
beam as soon as the power is turned on. Turn the power on after installing both the laser unit and
sensor control unit to the process flange or flow cell (in a condition where the laser beam is not
irradiated outside the process).
When the TDLS8000 is turned on, transmission is indicated on the display of the LU and that of
the SCU. The LU displays the value with 4 digits, and the SCU as “Trans **.*%.” The alignment
flange needs to be adjusted using the angle fine-adjustment nut so that this transmission is
maximized. (See Figure 3.16.)
The transmission display is updated every analysis period. The standard analysis period is 2 to 5
seconds. For details on the analysis period, see “Appendix 5 What Is an Analysis Period?”.
While adjusting the optical axis, check the updated display showing the most recent
transmission. Note that it is possible to determine when the transmission is updated by the way
each display changes its displayed content.
For details, see “1.2 Name and Function of Each Part”.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3-24
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
3.3.1
Optical Axis Adjustment When Not Using an LAO Unit
(Optical path length 6 m or less)
As shown in Figure 3.16, alignment flanges are attached to the LU and SCU. The angle
adjustment nuts of these flanges are used to adjust the optical axis.
Alignment flange
Angle adjustment nut (8)
Instrument mounting flange
Process flange
Quick connector
Bellows
Process flange
Alignment flange
LU or SCU
Figure 3.16
Optical axis adjustment when not using an LAO unit
Use the angle adjustment nuts of the alignment flange shown in Figure 3.16 to adjust the optical
axis. For the optical axis adjustment on the LU and SCU, check the transmission each time you
turn an angle adjustment nut quarter of a turn. Wait until the transmission shown on the display is
updated at least twice. Repeat this process while regularly checking the transmission value.
Follow the procedure below to adjust the optical axis. Adjust one side at a time starting with the
LU side. The proper optical axis adjustment may not be achieved when the optical axis on the LU
and SCU is changed at the same time.
(1) Checking the installation condition (LU and SCU)
Check the initial condition of the alignment flanges to see if the LU and SCU are facing each
other. If it is apparent that they are not, adjust the alignment coarsely before continuing with
the rest of the procedure.
(2) Coarsely adjusting the laser unit (LU)
Loosen the angle adjustment nuts on the LU side.
While viewing the transmission value, move the LU with your hand to perform coarse
adjustment. Repeat the adjustment in the vertical direction first and then the horizontal
direction.
Find the position that yields the maximum transmission value, and tighten the angle
adjustment nuts with your fingers. From this position, begin the fine adjustment.
If the transmission value does not increase when you coarsely adjust the LU, the mounting
position of the SCU may not be correct. If this occurs, loosen the angle adjustment nuts on
the SCU side, adjust it to a position that yields a higher transmission value, and then repeat
step (2).
If the transmission value does not change at all even when you make these adjustments,
check that the positions of the process flange on the LU and SCU sides are correct.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3-25
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
(3) Finely adjusting the laser unit (LU)
a. Vertical adjustment
Adjust the angle adjustment nuts while checking the transmission value to adjust the
optical axis in the vertical direction. Adjust the nut to the position that yields the maximum
transmission value.
b. Horizontal adjustment
Adjust the angle adjustment nuts while checking the transmission value to adjust the
optical axis in the horizontal direction. Adjust the nut to the position that yields the
maximum transmission value.
c. Fine adjustment
After adjusting the optical axis in the horizontal direction, make fine adjustments in the
vertical direction again. After that, make fine adjustments in the horizontal direction.
Repeat these final adjustments until the maximum transmission is obtained.
(4) Finely adjusting the sensor control unit (SCU) side
a. Vertical adjustment
Adjust the angle adjustment nuts while checking the transmission value to adjust the
optical axis in the vertical direction. Adjust the nut to the position that yields the maximum
transmission value.
b. Horizontal adjustment
Adjust the angle adjustment nuts while checking the transmission value to adjust the
optical axis in the horizontal direction. Adjust the nut to the position that yields the
maximum transmission value.
c. Fine adjustment
After adjusting the optical axis in the horizontal direction, make fine adjustments in the
vertical direction again. After that, make fine adjustments in the horizontal direction.
Repeat these final adjustments until the maximum transmission is obtained.
After the optical axis adjustment on the SCU side is complete, securely fasten the angle
adjustment nuts.
(5) Finely adjusting the laser unit (LU) side
Again, finely adjust the laser unit (LU) side. Repeat step (3) above until the transmission
value is finally at the maximum value. After the optical axis adjustment on the LU side is
complete, securely fasten the angle adjustment nuts.
(6) Checking the angle adjustment nuts are fixed in place
Check that the angle adjustment nuts on the LU side and SCU side are fixed securely in
place.
(7) Transmission calibration
3.3.2
After the above adjustment is complete, perform transmission calibration by assuming the
maximum transmission value that was obtained to be 100%. (See “9.1.1 Transmission
Calibration” for this procedure).
Optical Axis Adjustment When Using an LAO Unit
(Optical path length 6 m or more)
For applications whose optical path length is 6 m or longer, select the “-LA” TDLS8000 optics
accessory. The emitted laser beam is adjusted so that it spreads out slightly, and a condensing
lens unit (LAO unit) is added to the SCU side. On the SCU side, the projected area of the laser
beam spreads out making it easy to adjust the optical axis to obtain stable transmission.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3-26
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
An alignment flange as shown in Figure 3.16 is mounted on the LU, and an LAO unit as shown in
Figure 3.17 is mounted on the SCU. In all cases, the angle adjustment nuts are used to adjust the
optical axis. Although the LAO unit is used, the laser beam position will vary greatly even with a
slight change in the optical axis because the process optical path length is long. As such, a more
detailed adjustment will be necessary.
Alignment flange
Angle adjustment nut (8)
Instrument mounting flange
Process flange
Nut (8)
Quick connector
Bellows
Alignment part
SCU
Figure 3.17
Process flange
LAO unit
Optical axis adjustment when using an LAO unit (SCU side)
Use the angle adjustment nuts shown in Figure 3.17 to adjust the optical axis. Check the
transmission each time you turn an angle adjustment nut one-eighth of a turn. Wait until the
transmission shown on the display is updated at least twice. Repeat this process while regularly
checking the transmission value.
Follow the procedure below to adjust the optical axis when the LAO is used. Adjust one side at a
time starting with the LU side. The proper optical axis adjustment may not be achieved when the
optical axis on the LU and SCU is changed at the same time.
(1) Checking the installation condition (LU and SCU)
Check the initial condition of the alignment flange and the LAO unit to see if the LU and SCU
are facing each other. If it is apparent that they are not, adjust the alignment coarsely before
continuing with the rest of the procedure.
(2) Coarsely adjusting the laser unit (LU) side
Loosen the angle adjustment nuts on the LU side.
While viewing the transmission value, move the LU with your hand to perform coarse
adjustment. Repeat the adjustment in the vertical direction first and then the horizontal
direction.
Find the position that yields the maximum transmission value, and tighten the angle
adjustment nuts with your fingers. From this position, begin the fine adjustment.
While making adjustments, keep in mind that the change in transmission may be extremely
small depending on the laser incident angle to the SCU. Not paying attention to this fact may
hinder the correct optical axis adjustment of the SCU.
If change in the transmission cannot be detected, adjust both the vertical and horizontal
direction in more detail.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3-27
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
If the transmission value does not increase when you coarsely adjust the LU, the mounting
position of the SCU may not be correct. If this occurs, loosen the angle adjustment nuts
on the SCU side, adjust it to a position that yields a higher transmission value, and then
repeat step (2). If the transmission value does not change at all even when you make these
adjustments, double-check that the positions of the process flange on the LU and SCU
sides are correct.
(3) Coarsely adjusting the sensor control unit (SCU) side
If change in the transmission is detected, coarsely adjust the SCU to maximize the
transmission.
Loosen the angle adjustment nuts on the SCU side. To adjust the optical axis of the SCU,
slowly tighten the angle adjustment nuts one-eighth of a turn at a time until the maximum
transmission is obtained. Then fix the optical axis adjustment position.
(4) Finely adjusting the laser unit (LU)
a. Vertical adjustment
Adjust the angle adjustment nuts while checking the transmission value to adjust the
optical axis in the vertical direction. Adjust the nut to the position that yields the maximum
transmission value.
b. Horizontal adjustment
Adjust the angle adjustment nuts while checking the transmission value to adjust the
optical axis in the horizontal direction. Adjust the nut to the position that yields the
maximum transmission value.
c. Fine adjustment
After adjusting the optical axis in the horizontal direction, make fine adjustments in the
vertical direction again. After that, make fine adjustments in the horizontal direction.
Repeat these final adjustments until the maximum transmission is obtained.
(5) Adjusting the laser unit (LU) side one last time
Make a final adjustment so that the laser beam hits the center of the SCU with maximum
transmission.
a. Final vertical adjustment
!!!Adjust only the top and bottom angle adjustment nuts.!!!
i. Adjust the vertical angle adjustment nuts on the LU side so that the transmission is
nearly zero (so that the laser beam barely misses the SCU lens) (Figure 3.18(a)).
ii. Adjust the vertical angle adjustment nuts on the LU side in the direction opposite to
the direction of (a) until the transmission is nearly zero (until that the laser beam barely
misses the SCU lens) (Figure 3.18(b)). During this process, turn the angle adjustment
nuts one-eighth of a turn at a time and record how many one-eighth turns the nuts
were turned.
iii. Adjust the vertical angle adjustment nuts on the LU side in the direction opposite to
the direction of (b) until the transmission is nearly zero (until that the laser beam barely
misses the SCU lens) (Figure 3.18(a’)). During this process, turn the angle adjustment
nuts one-eighth of a turn at a time and record how many one-eighth turns the nuts
were turned.
iv. Check that the number of rotations of the angle adjustment nuts are nearly the
same for ii and iii. Then, return the vertical angle adjustment nuts half the number of
rotations, and securely fasten the nuts (Figure 3.18(c)).
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
3-28
(a) (a’)
SCU lens
(c)
(b)
Figure 3.18
Final vertical adjustment conceptual illustration
b. Final horizontal adjustment
Adjust only the left and right angle adjustment nuts.
i. Adjust the horizontal angle adjustment nuts on the LU side so that the transmission is
nearly zero (so that the laser beam barely misses the SCU lens) (Figure 3.19(d)).
ii. Adjust the horizontal angle adjustment nuts on the LU side in the direction opposite to
the direction of i until the transmission is nearly zero (until that the laser beam barely
misses the SCU lens) (Figure 3.19(e)). During this process, turn the angle adjustment
nuts one-eighth of a turn at a time and record how many one-eighth turns the nuts
were turned.
iii. Again, adjust the horizontal angle adjustment nuts on the LU side in the direction
opposite to the direction of ii until the transmission is nearly zero (until that the laser
beam barely misses the SCU lens) (Figure 3.19(d’)). During this process, turn the
angle adjustment nuts one-eighth of a turn at a time and record how many one-eighth
turns the nuts were turned.
iv. Check that the number of rotations of the angle adjustment nuts are nearly the same
for ii and iii. Then, return the horizontal angle adjustment nuts half the number of
rotations, and securely fasten the nuts (Figure 3.19(f)).
SCU lens
(e)
Figure 3.19
(c)
(f)
(d)
(d’)
Final horizontal adjustment conceptual illustration
(6) Coarsely adjusting the sensor control unit (SCU) side one last time
Make final vertical and horizontal adjustments to maximize the transmission. Repeat these
final adjustments until the maximum transmission is obtained. After the final adjustment on
the SCU side is complete, securely fasten the angle adjustment nuts.
(7) Checking the angle adjustment nuts are fixed in place
Check that the angle adjustment nuts on the LU side and SCU side are fixed securely in
place.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3-29
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
(8) Transmission calibration
After the above adjustment is complete, perform transmission calibration by assuming the
maximum transmission value that was obtained to be 100%. (See “9.1.1 Transmission
Calibration” for this procedure).
3.4 Piping
After wiring and optical axis adjustment are complete, connect the pipes for the purge gas.
After piping is complete, to keep the TDLS8000 process window area clean, we recommend that
you let the purge gas flow until the beginning of operation.
CAUTION
To maintain the dust proof and waterproof performance of the TDLS8000, attach pipes or plugs
to all ports.
For the piping thread specifications, check the inscriptions near the ports. See Figure 3.10 for an
example.
n Piping Parts
Refer to the following table, and check that all the necessary piping parts are available.
Device
Piping location
Purge port
TDLS8000
Alignment flange purge port
LAO purge port
YC8000
Piping port
IF8000
Purge port
Calibration cell Piping port
Piping parts
Tube joint
Tube joint
Tube joint
Tube
Tube joint
Tube joint
Remarks
Rc1/4 or 1/4NPT, off-the-shelf product
Rc1/4 or 1/4NPT, off-the-shelf product
Rc1/4 or 1/4NPT, off-the-shelf product
1/4 inch stainless pipe, off-the-shelf product
Rc1/4 or 1/4NPT, off-the-shelf product
1/4NPT, off-the-shelf product
n Purge Gas
Refer to the information provided in the specifications of chapter 2.
(1) Purge gas type
Normally, nitrogen (N2) is used for the purge gas, but depending on the application, instrumental
air may suffice.
Use nitrogen gas or instrumental air that meets the following conditions.
• Is clean. Dust particle diameter is less than 0.5 µm.
• Does not contain oil.
• Nitrogen gas with 99.99% or higher purity when measuring O2 or ppmH2O.
• Nitrogen gas with less than 20 ppm moisture content when measuring ppmH2O.
(2) Purge gas flow rates
Feed purge gases with the following purge flow rates.
• Analyzer internal purge, validation purge:
2 to 10 L/min (depending on the application)
150 mL/min for internal purge and 2 to 10 L/min for validation purge (for Div 1, Zone 1 Type)
• Process window purge:
5 to 300 L/min (depending on the application)
Insert a flowmeter with an adjustable needle valve between the supply line and purge inlet when
feeding purge gas.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3-30
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
(3) Exhausting purge gas
Purge gas exhaust is as follows.
• Analyzer internal purge, validation purge:
Connect pipes to outlet ports if necessary to exhaust the purge gas to an
appropriate location. Construct them so that rainwater and the like do not enter
the ports.
If you are using hazardous gas (e.g., CO gas) for check gas, exhaust it inside the
process or in an appropriate manner.
• Process window purge:
The gas is exhausted inside the process.
n Areas That Need to Be Purged
The TDLS8000 needs to be purged with nitrogen gas for the following two purposes.
First is the continuous nitrogen gas purging to prevent oxygen from open air and moisture from
entering the measurement optical path during oxygen concentration, moisture concentration, and
other process gas concentration measurements. This is called analyzer internal purge.
To perform validation, areas under analyzer internal purge is temporarily replaced with check
gas. This is called validation purge.
Analyzer internal purge and validation purge are applied to the following five areas (see Figure
3.20).
(1) Laser module area (inside the LU)
(2) SCU module area (inside the SCU)
(3) Validation area (inside the SCU)
(4) LU alignment flange area (LU side: when a process isolation flange is used)
(5) SCU alignment flange area (SCU side: when a process isolation flange is used)
Figure 3.20
Areas to be purged
The second is the continuous purge on the process window (the interface to the process gas) to
keep it clean (prevent the process from the adherence of dust). This is called process window
purge. In process window purge, purge gas is fed inside the process from the purge port closest
to the process side.
NOTE
For Explosionproof/Flameproof type, refer to “3.4.3 Purge Gas Piping for Explosionproof/
Flameproof type”.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3.4.1
3-31
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
Purge Gas Piping for In-situ Installation
(1) When not using the validation function
If the installation into the process is in situ and the validation function is not used, connect the
piping as shown in Figure 3.21. For analyzer internal purge, feed nitrogen gas through purge port
A in Figure 3.21 for both the LU and SCU. For process window purge, feed nitrogen gas through
the purge port of the alignment flange.
Process flange
N2 gas
LU side
Exhaust
N2 gas
SCU side
Exhaust
Measured gas
A
A
Alignment flange
N2 gas
Sealing plug
Figure 3.21
Alignment flange
Sealing plug
N2 gas
Piping for in-situ installation when not using the validation function
(2) When using the validation function
If the installation into the process is in situ and the validation function is used, connect the piping
as shown in Figure 3.22. For analyzer internal purge, feed nitrogen gas through purge port A
in Figure 3.22 for both the LU and SCU. For the SCU, since check gas needs to be fed during
validation, connect the piping with a three-way valve so that the gas can be switched between
nitrogen gas and check gas. For process window purge, feed nitrogen gas through the purge port
of the alignment flange in the same manner as “(1) When using the validation function.”
Process flange
N2 gas
LU side
Exhaust
N2 gas
SCU side
Exhaust
Measured gas
A
N2 gas
A
Alignment flange
Sealing plug
Alignment flange
Sealing plug
Switching valve
N2 gas
Checking gas
Figure 3.22
Piping for in-situ installation when using the validation function
(3) When using a process isolation flange but not the validation function
If the installation (in-situ or bypass) into the process uses a process isolation flange but validation
will not be conducted, connect the piping as shown in Figure 3.23.
For analyzer internal purge, feed nitrogen gas through purge port A in Figure 3.23 for both the LU
and SCU. Connect the piping so that the TDLS8000 purge gas also flows through the inside of
the alignment flange.
For process window purge, feed nitrogen gas through the purge port of the process isolation
flange.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3-32
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
LU side
Sealing plug
Alignment flange
Exhaust
Process flange
Alignment flange
Exhaust
Sealing plug
SCU side
Measured gas
A
A
Sealing plug
N2 gas
Process insulation flange
Sealing plug
N2 gas
N2 gas
N2 gas
Process insulation flange
Figure 3.23
Piping when using a process isolation flange (in-situ or bypass installation) but not the
validation function
(4) When using a process isolation flange and the validation function
If the installation (in-situ or bypass) into the process uses a process isolation flange and
validation will be conducted, connect the piping as shown in Figure 3.24. For analyzer internal
purge, feed nitrogen gas through purge port A in Figure 3.24 for both the LU and SCU. For the
SCU, since check gas needs to be fed during validation, connect the piping with a three-way
valve so that the gas can be switched between nitrogen gas and check gas. For process window
purge, feed nitrogen gas through the purge port of the process isolation flange in the same
manner as “(3) When using a process isolation flange and the validation function.”
LU side
Sealing plug
Alignment flange
Exhaust
Process flange
Alignment flange
Exhaust
Sealing plug
SCU side
Measured gas
A
N2 gas
A
Sealing plug
Process insulation flange
Sealing plug
N2 gas
N2 gas Process
insulation
flange
Switching valve
N2 gas
Checking gas
Figure 3.24
Piping when using a process isolation flange (in-situ or bypass installation) and the
validation function
3.4.2
Purge Gas Piping for Sampling System Using Flow
Cells
(1) When not using the validation function
If the application uses a flow cell but not the validation function, connect the piping as shown in
Figure 3.25.
The piping for analyzer internal purge is the same as “(3) When using a process isolation flange
but not the validation function.” Feed nitrogen gas through purge port A in Figure 3.25 for both the
LU and SCU. Connect the piping so that the TDLS8000 purge gas also flows through the inside
of the alignment flange.
Process window purge is not used. The flow cell is purged directly with the process gas.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3-33
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
LU side
Sealing plug
Exhaust
Exhaust
Flow cell
Sealing plug
A
N2 gas
A
Sealing plug
Figure 3.25
SCU side
Process gas
Exhaust
N2 gas
Sealing plug
Piping when not using the validation function
(2) When using the validation function
If the application uses a flow cell and the validation function, connect the piping as shown in
Figure 3.26.
The piping for analyzer internal purge is the same as “(4) When using a process isolation flange
and the validation function.” Feed nitrogen gas through purge port A in Figure 3.26 for both the
LU and SCU. For the SCU, since check gas needs to be fed during validation, connect the piping
with a three-way valve so that the gas can be switched between nitrogen gas and check gas.
Process window purge is not used. The flow cell is purged directly with the process gas.
LU side
Sealing plug
Exhaust
Exhaust
Flow cell
Sealing plug
A
N2 gas
SCU side
A
Sealing plug
Process gas
Exhaust
Sealing plug
Switching valve
N2 gas
Checking gas
Figure 3.26
3.4.3
Piping when using the validation function
Purge Gas Piping for Explosionproof/Flameproof type
In Explosionproof/Flameproof type TDLS8000 (TDLS8000-C1, D1, E1 and S1), analyzer internal
purge area, that is laser module area and detector module area, described in 3.4 are divided
into two areas by explosion protection window as shown Figure 3.27. When using Explosion
proof/Flame proof type, piping for purge gas should be connected to each area separately. For
validation, SCU area and LU area are used.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3-34
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
Analytical internal purge area
SCU Ex area
SCU area
Explosion protection connector
LU area
LU Ex area
Explosion protection connector
Explosion protection window
Figure 3.27
Areas to be purged in Explosion proof/Flame proof type
(1) Explosionproof/Flameproof type without validation function
If TDLS8000 Explosionproof/Flameproof type is installed in situ and the validation function is not
used, connect piping as shown in Figure 3.28. For SCU Ex area and LU Ex area, feed nitrogen
gas through purge port A at 150 mL/min and for SCU and LU area, feed nitrogen gas through
purge port B in Figure 3.28. The piping for these two areas in a unit should be separated because
required flow rate is different. For flow rate of SCU and LU are, refer to 3.4 (2). For process
window purge, feed nitrogen gas through the purge ports of the alignment flange.
N2 gas
LU side
Process flange
Exhaust
N2 gas
N2 gas
Exhaust
N2 gas
SCU side
Measured gas
A
N2 gas
(150 mL/min)
Figure 3.28
A
Exhaust
Alignment flange
Alignment flange
N2 gas
Exhaust (150 mL/min)
Piping for Explosion proof/Flame proof type without validation
CAUTION
Excess flow rate for SCU and LU Ex area may cause Detector or Laser module damage.
WARNING
After stoppage of purge gas, explosive gas may be inside of the enclosures. Purge inside of the
each enclosure, SCU Ex area and LU Ex area, using nitrogen for at least 60 minutes at 10 kPa at
the inlet of the enclosure before turning on the power to prevent from pressurizing internal gas by
purge gas. See Appendix 8.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
3-35
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
(2) Explosionproof/Flameproof type with validation function
If TDLS8000 is installed in the process is in situ and the validation function is used, connect the
piping as shown in Figure 3.29. For SCU Ex area and LU Ex area, feed nitrogen gas through
purge port A at 150 mL/min and for SCU area and SCU and LU area, feed nitrogen gas through
purge port B in Figure 3.29. The piping for these two areas in a unit should be separated because
flow rate is different. For flow rate of SCU and LU are, refer to 3.4 (2). For process window purge,
feed nitrogen gas through the purge ports of the alignment flange. For validation, SCU area and
LU area is used as validation cell and connect the piping with a three-way valve so that the gas
can be switched between nitrogen gas and check gas. For process window purge, feed nitrogen
gas through the purge port of the alignment flange.
Process flange
N2 gas
LU side
Exhaust
N2 gas
SCU side
Exhaust
Measured gas
B
B
A
N2 gas
(150 mL/min)
A
Exhaust
Alignment flange
Alignment flange
N2 gas
(150 mL/min)
Switching valve
N2 gas
Exhaust
Checking gas
Figure 3.29
Piping for Explosion proof/Flame proof type with validation
CAUTION
• Excess flow rate for SCU and LU Ex area may cause Detector or Laser module damage.
• Validation gas line cannot be connected to SCU or LU Ex area.
WARNING
After stoppage of purge gas, explosive gas may be inside of the enclosures. Purge inside of the
each enclosure, SCU Ex area and LU Ex area, using nitrogen for at least 60 minutes at 10 kPa at
the inlet of the enclosure before turning on the power to prevent from pressurizing internal gas by
purge gas. See Appendix 8.
(3) Explosionproof/Flameproof type with Flow cell or Isolation flange
If TDLS8000 is installed with an isolation flange or a flow cell, refer to this clause and 3.4.1 (3),
(4) or 3.4.2. The piping for analyzer internal purge area is the same as 3.4.3 (1) or (2) except the
outlets of SCU and LU area adjacent each alignment flange. Those outlets should be connected
to the inlets of alignment flange. The piping for the flow cell or isolation flange is the same as the
piping described in 3.4.1 (3), (4) or 3.4.2.
CAUTION
• Excess flow rate for SCU and LU Ex area may cause Detector or Laser module damage.
• Validation gas line cannot be connected to SCU or LU Ex area.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
3-36
WARNING
After stoppage of purge gas, explosive gas may be inside of the enclosures. Purge inside of the
each enclosure, SCU Ex area and LU Ex area, using nitrogen for at least 60 minutes at 10 kPa at
the inlet of the enclosure before turning on the power to prevent from pressurizing internal gas by
purge gas. See Appendix 8.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
4.
4-1
<4. YH8000 Installation>
YH8000 Installation
There are two methods of YH8000 HMI unit installation: local HMI installation in which the unit is
mounted directly on the TDLS8000 and remote HMI installation in which the unit is mounted by
itself on a wall or the like.
See “2.3
Model and Codes”.
CAUTION
During installation, be careful not to drop the product, damage the display, and so on.
l Installation location
The YH8000 is designed to work even in harsh environmental conditions, but note the following
points in order to use it stably for a long time.
Select an installation location where the ambient temperature and humidity are within the
specifications.
Be sure to block direct sunlight such as by attaching a sun shield or installing it in a cubicle.
If the YH8000 it subject to heat radiation such as from plant facilities, take heat insulation
measures.
Also, select a location that meets the following conditions.
• Where there is hardly any mechanical vibration or shock
• Where it is not subject to direct sunlight or harsh weather conditions
• Where there is no corrosive atmosphere
4.1
Local HMI Installation
The YH8000 can be mounted on the TDLS8000 by using the mounting kit for TDLS8000 (option
code: /M). The YH8000 can be mounted on the front, right, or left.
If you want the front cover to open in the opposite direction due to the installation location, the
YH8000 can be mounted upside down.
The sun shield (option code: /S) for avoiding the effects of direct sunlight can also be mounted.
When attaching the mounting kit to the YH8000, fasten the supplied TDLS8000 bolt and YH8000
bolt with a torque of about 5 to 6 N•m.
l Procedure for attaching the bracket to the front
(1) Fasten accessory brackets 1 and 2 in place with the YH8000 bolts (M6x14 mm).
You can tilt the YH8000 in the range of -20° to 20° by changing the angle adjustment hole
used to fasten accessory bracket 1 in place.
(2) Fasten accessory bracket 1 to the YH8000 with the YH8000 bolts.
(3) Fasten accessory bracket 3 to the TDLS8000 with the TDLS8000 bolts (M6x10 mm).
Note that the length of the TDLS8000 bolts and YH8000 bolts is different.
(4) Fasten accessory bracket 2 to accessory bracket 3 with the YH8000 bolts.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
4-2
<4. YH8000 Installation>
Accessory bracket 1
Washer
YH8000 bolt
(with a hexagonal hole, M6 x 14 mm)
Accessory bracket 2
Angle adjustment hole
Accessory bracket 3
TDLS8000 bolt
(with a hexagonal hole, M6 x 10 mm)
Figure 4.1
Mounting to the front
l Procedure for attaching the bracket to the side
(1) Fasten accessory brackets 1 and 2 in place with the YH8000 bolts.
You can tilt the YH8000 in the range of -20° to 20° by changing the angle adjustment hole
used to fasten accessory bracket 1 in place.
(2) Fasten accessory bracket 1 to the YH8000 with the YH8000 bolts.
(3) Fasten accessory bracket 3 to the TDLS8000 with the TDLS8000 bolts.
(4) Fasten accessory bracket 2 to accessory bracket 3 with the YH8000 bolts. You can fasten it
to the left or the right side.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
4-3
<4. YH8000 Installation>
Accessory bracket 1
Washer
YH8000 bolt
(with a hexagonal hole, M6 x 14 mm)
Accessory bracket 2
Angle adjustment
hole
Accessory bracket 3
TDLS8000 bolt
(with a hexagonal hole, M6 x 10 mm)
Figure 4.2
Mounting to the right
l When attaching the sun shield
Place the sun shield over the YH8000, and then fasten accessory bracket 1 in place with the
YH8000 bolts. Fasten the bolts with a torque of about 5 to 6 N•m.
Sun shield
Figure 4.3
Attaching the sun shield
l Inverted YH8000 installation
The YH8000 can be mounted upside down so that the cables come out on the left side. When
you mount the YH8000 upside down, you can invert the screen using the inversion switch. For
details on how to use the inversion switch, see page 4-3 of “l Inverted YH8000 installation”.
This mounting method can be used when mounting the YH8000 on the mounting kit for
TDLS8000, a wall, or a pipe. For details on the mounting bracket and the YH8000 tightening
torque, see the details of each mounting method.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
Figure 4.4
4.2
4-4
<4. YH8000 Installation>
Upside down mounting
Wiring for Local HMI Installation
After mounting the YH8000 on the TDLS8000, wire it.
If the YH8000 is mounted upside down, reverse the YH8000 inversion switch.
CAUTION
Never turn on the power to the TDLS8000 or the devices connected to the TDLS8000 until all
wiring is complete.
l Opening the YH8000 front cover
Open the front cover according to the following procedure.
Front cover fastening screw
How to open the front cover
(1) Loosen the M5 screws holding the front cover. The screws are designed not to come off the
front cover.
(2) Open the front cover outward to the left.
How to close the front cover
(1) Close the front cover. Be careful not to get your fingers or other body parts caught in the
hinge area and between the front cover and the case.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
4-5
<4. YH8000 Installation>
(2) Align the front cover to the screw holes of the case, and fasten the screws evenly. Do not
tighten any screw all the way. When the gasket load starts to be applied to the four corner
screws, tighten the screws evenly about one turn. Use a tightening torque of 1.8 to 2.0 N•m.
CAUTION
When opening or closing the front cover, be careful not to get your fingers or other body parts
from being caught between the front cover and the case.
CAUTION
When opening the front cover, check that the screws are lifted completely off the screw holes,
and open the cover slowly. This is to prevent damaging the threaded parts of the housing.
If the threaded parts are damaged and the screws cannot be tightened securely, waterproof
capability will be lost.
l Cable entries
There are two cable entries on the YH8000. Attach conduits or cable glands with the appropriate
thread size to the YH8000 cable entries.
Symbols are inscribed near the cable entries for identifying the thread specifications.
For ANSI 1/2NPT:
A
For ISO M20x1.5:
M
Cable entry 2
1/2NPT or M20
Covered with a plug
Cable entry 1
1/2NPT or M20
Cable entry 1
Cable entry for the cable that connects to the TDLS8000.
Cable entry 2
A plug with a hexagonal hole covers this entry. To use Port2 for Modbus
communication, remove the plug and run the cable through this entry. This entry
is also used when running an Ethernet cable for remote HMI installation.
Be careful of the tightening torque when assembling cable glands.
l Wiring
CAUTION
Be careful not to connect the power supply wires to the incorrect locations or reverse the polarity.
Incorrectly connecting the power supply can cause a malfunction.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
4-6
<4. YH8000 Installation>
POWER
-
+
Terminal
Function
name
POWER+ YH8000 power supply input. 24 V DC±10%
POWER- In the case of a local connection, power is supplied from the SCU.
In the case of special cables, connect two to the positive terminal and two
to the negative.
PORT1 Connection to the SCU
PORT2 Modbus communication, Ethernet communication for remote connection
For the cables you need to use, see page 3-13 of “n
Types of Wiring and Cabling”.
When the YH8000 is mounted on the TDLS8000 (local HMI installation), you need to use a local
HMI connection cable (option code: /C), which is a special option cable. The wiring of the special
cable is explained below.
This cable must be processed before it is connected. For details on how to process the cable,
see “Appendix 2
Constructing Local HMI Connection Cables”.
(1) Connect the cable to the power supply terminal with the correct polarity. Use a tightening
torque of 0.22 to 0.25 N•m for the power supply terminal.
(2) Insert the RJ45 connector into Port1 until you hear a click.
(3) Connect the shielded wire of the special cable to the ground terminal for the shielded wire.
(4) Connect a grounding cable that complies with page 3-13 of “n Types of Wiring and Cabling”
to the YH8000 ground terminal.
Port 2
Port 1
Ground terminal
for shielded wire
Power supply
terminal
Ground
terminal
l Setting the YH8000 inversion switch
NOTE
Set the inversion switch with the YH8000 turned off. Turning it off applies the setting.
The YH8000 can be positioned so that the cables come out on the right side or the left side.
Depending on the direction, you need to reverse the YH8000 inversion switch, which is inside the
case. The inversion switch is for inverting the screen upside down. Set the switch according to
how the YH8000 is mounted as shown below.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
4-7
<4. YH8000 Installation>
The factory default setting is for the cables to come out from the right side.
INV
NORM
YH8000 inversion switch
YH8000 inversion switch
How the YH8000 is mounted
Cables coming out from the right side.
Cables coming out from the left side.
4.3
Inversion switch position
NORM
INV
Remote HMI Installation
This section explains how to mount the YH8000 using mounting brackets (pipe mount (option
code: /P), wall mount (option code: /W)).
The sun shield (option code: /S) for avoiding the effects of direct sunlight can also be mounted.
WARNING
The TDLS8000 does not have a power switch. Provide a switch on the power supply line to
separate the TDLS8000 from the main power supply. Use labels to indicate that this switch is for
cutting off the power supply to the TDLS8000 and to indicate ON and OFF.
l Pipe mounting
The YH8000 can be mounted on a 50A (2B) pipe. Fasten the YH8000 bolts with a torque of about
5 to 6 N•m. The YH8000 can be mounted horizontally or vertically.
The pipe mount option includes short bolts that are used when a sun shield is used and long bolts
that are used when a sun shield is not used.
Washer
Long bolt for
pipe mounting
Vertical mounting
Figure 4.5
Horizontal mounting
Pipe mounting
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
4-8
<4. YH8000 Installation>
When attaching a sun shield, spacers are used to fasten the sun shield to the YH8000 before
mounting on the pipe.
Spacer
Washer
Short bolt for
pipe mounting
Figure 4.6
Pipe mounting when a sun shield is used (vertical mount example)
l Wall mounting
For the wall mount dimensions, see “2.3 Model and Codes”.
Fasten the wall mount bracket to the YH8000 with a tightening torque of about 5 to 6 N•m.
M6 bolt mounting hole (4 locations)
* Bolts are not included.
Washer
Bolt for wall mounting
(M6 bolt with a hexagonal hole)
Figure 4.7
Wall mounting
When attaching a sun shield, turn the wall mount bracket so that the longer side is vertical.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
4-9
<4. YH8000 Installation>
M6 bolt mounting hole (4 locations)
* Bolts are not included.
Washer
Bolt for wall mounting
(M6 bolt with a hexagonal hole)
Figure 4.8
4.4
Wall mounting when a sun shield is used
Wiring for Remote HMI Installation
Wire the cables after completing the remote HMI installation.
CAUTION
Never turn on the power to the YH8000 or the devices connected to the YH8000 until all wiring is
complete.
l Opening the YH8000 front cover
See “4.2
Wiring for Local HMI Installation”.
l Cable entries
Run the power cable through cable entry 1 shown in “4.2
Wiring for Local HMI
Installation”. Run the Ethernet cable through cable entry 2 after removing the plug with a
hexagonal hole.
l Wiring
See “4.2
Wiring for Local HMI Installation”.
CAUTION
Be careful not to connect the power supply wires to the incorrect locations or reverse the polarity.
Incorrectly connecting the power supply can cause a malfunction.
For the cables you need to use, see page 3-13 of “n
Types of Wiring and Cabling”.
Use two-core or three-core shielded cable to wire the power supply.
If you want to connect the functional ground terminal inside the instrument, use a three-core
power cable.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<4. YH8000 Installation>
4-10
l Connecting an Ethernet Cable
If you want to connect the YH8000 to TDLS8000 through an Ethernet hub, you will need to
use an Ethernet cable. The Ethernet cable connector must be crimped during the YH8000
installation.
The processing of the Ethernet cable is the same as for the TDLS8000. See “3.2.8 Connecting
an Ethernet Cable”.
Power supply wiring, ground wiring, and Ethernet cable connection when using the YH8000
remotely are explained below. For the terminal positions, see the wiring diagram in “4.2 Wiring
for Local HMI Installation”.
(1) Connect the cable to the power supply terminal with the correct polarity. Use a tightening
torque of 0.22 to 0.25 N•m for the power supply terminal.
(2) Insert the RJ45 connector into Ethernet Port2 until you hear a click.
(3) Connect the shielded wire of the special cable to the ground terminal for the shielded wire.
(4) Connect a grounding cable that complies with page 3-13 of “n Types of Wiring and Cabling”
to the YH8000 ground terminal.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
5-1
<5. Startup>
5. Startup
Refer to “3.2 Wiring” and “3.4 Piping”, and verify that the system has been constructed
correctly. Perform the startup procedure with the optical axis adjustment completed.
Run analyzer internal purge gas and process window purge gas at the appropriate flow rates.
Supply power to the TDLS8000.
The SCU display shows a screen indicated in “1.2 Name and Function of Each Part”. The LU
display shows the transmission value [%].
NOTE
Even in an application that requires the system to be regularly run and suspended repeatedly,
we recommend that you continuously supply power and process window purge gas to the
TDLS8000. This is to prevent unnecessary temperature changes in and placing unnecessary
load on the laser device and sensor.
5.1
Connecting the HART Configuration Tool
This section explains how to connect the HART configuration tool and provides a brief overview
of the menu tree shown on the tool. For details on the menu tree and HART communication
function, see “7. HART Communication” and “Appendix 3 General View of HART DD”.
NOTE
Write protection based on password authentication is available for TDLS8000 HART
communication. By factory default, the write protection is disabled. Therefore, you can change
the settings simply by connecting the configuration tool. For details on how to enable write
protection, see “7. HART Communication”.
5.1.1
Installing a DD File
Before you can start using the HART configuration tool, the TDLS8000 DD (Device Description)
must be installed in the configuration tool. If you want to use FieldMate for the configuration
tool, obtain the latest Device Files, and install a DTM. For details, see the FieldMate instruction
manual.
The following table shows the relation between DD and Device Files.
DD Revision Device Files Revision
01
3.06.11
02
3.06.24
If you want to connect your own configuration tool, download the DD file from the YOKOGAWA
website and install it.
5.1.2
https://partner.yokogawa.com/global/interoperability/dd-file-hart_an.htm
Connection Procedure
Connect the configuration tool in parallel with the load resistance connected to the analog output
AO-1 terminal. There is no polarity. For details on connecting the load resistance, see “3.2
Wiring”. Figure 5.1 shows a wiring example.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
5-2
<5. Startup>
TDLS8000 SCU
AO-1 +
4-20 mA DC
Relaying
terminals
Relaying
terminals
Control room
AO-1 USB
FieldMate
modem
USB
FieldMate
modem
PC/FieldMate
Figure 5.1
5.1.3
Receiving resistance
(Include cable resistance):
250Ω to 550Ω
PC/FieldMate
HART configuration tool wiring example
Basic Menu Configuration
The root menu of the menu tree displayed in the HART configuration tool is described below. For
the entire menu structure, see “7.2
Menu Tree”. For the entire menu including parameter
names, see “Appendix 3
General View of HART DD”.
Root menu
Process variables
Diagnosis/Service
Basic setup
Detailed setup
Review
Description
Displays the most recent PV-QV and measured values
Checks alarms and history and executes calibration, validation, and loop
checks
Assigns PV-QV items and sets output ranges
Sets TDLS8000-specific parameters
Displays measured values, I/O values, and production information
On FieldMate, the top menu structure is different from the DD menu. The FieldMate root menu
is “DTM Menu(Online),” and under it are the five DD root menus shown above. The lower level
structure, however, is the same as DD.
DTM Menu(Online)
• Device
Configuration
• Device setup
• Diagnostic
• Diagnostic/Service
• Process Variable
• Process variable
• Maintenance
• Maintenance
• Basic setup
• Detailed setup
• Review
(*1)
*1: Of the commands in the “Diagnostic/Service” menu, the execution commands for calibration, validation, and loop check
are included.
DTM root menu
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
5-3
<5. Startup>
NOTE
If you are using the Field Communicator 475 configuration tool, the following screen may appear
depending on the factory default settings of the TDLS8000. This is a notification that the same
measured value was read multiple times within the same analysis period. Since this is not
a problem with the TDLS8000 operation, select “YES.” This screen will appear every time a
connection is made. The analysis period is a fixed adjustment value assigned to each TDLS8000
and cannot be changed.
5.2
Connecting to the YH8000
This section describes how to connect the YH8000 HMI unit to the TDLS8000 and provides a
basic description of the related screens. For a detailed explanation of the YH8000, see “8.
YH8000 HMI Unit”.
5.2.1
Initialization and Connection Procedure
If you are connecting the TDLS8000 and YH8000 in a one-to-one configuration, the factory
default settings can be used. The settings need to be changed when you are connecting the
TDLS8000 to an existing LAN network or when you are connecting multiple TDLS8000s or
YH8000s.
The initial IP address settings are shown in Figure 5.1.
Table 5.1
Model
Factory default IP address settings
IP address
TDLS8000 192.168.1.10
YH8000
192.168.1.100
Subnet mask
255.255.255.0
255.255.255.0
Default gateway
192.168.1.254
192.168.1.254
To change the TDLS8000 IP address and subnet mask from the YH8000, follow the procedure
below.
(1) Connect the YH8000 using the factory default IP settings.
(2) Change the TDLS8000 IP address and subnet mask (see section 5.2.2).
(3) To change the IP address of multiple TDLS8000s, connect to the TDLS8000s one at a time
and change the IP address and subnet mask.
(4) If necessary, change the YH8000 IP address and subnet mask (see section 5.2.2).
(5) Change the YH8000 destination IP address (see section 5.2.3).
The procedure from turning on the YH8000 to connecting it in step (1) above is provided
below. The procedure assumes that the TDLS8000 has the factory default IP address settings.
To connect to a TDLS8000 whose IP address has been changed or to connect to multiple
TDLS8000s, see “5.2.3 Connecting to the TDLS8000”.
(1) Complete the wiring of the TDLS8000 and YH8000 according to the instructions in “3.2.8
Connecting an Ethernet Cable”.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
5-4
<5. Startup>
(2) Turn on the TDLS8000 and YH8000.
=> The opening screen (a) is displayed for about 10 seconds.
=> If they do not connect automatically, the YH8000 configuration screen (b) will appear. => Step (3)
=> If they connect automatically, the home screen (d) will appear.
(3) Tap Analyzer Connection to switch to the TDLS8000 selection screen (c). => Step (4)
(4) Tap Connect to start a connection.
=> If a connection is established, the home screen (d) will appear.
a
b
c
d
5.2.2
Setting the IP Address
l Setting the TDLS8000 IP address
Configuration menu path:
“
>>Analyzer>>Configuration>>System>>Communication>>TCP/IP”
NOTE
When you change the TDLS8000 IP address, the TDLS8000 automatically restarts with the new
IP address. The connection to the YH8000 will be disconnected. Follow the procedure below to
reconnect.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
5-5
<5. Startup>
The procedure for changing the TDLS8000 IP address and reconnecting is as follows. If you
change the TDLS8000 subnet mask, you will also need to change the YH8000 subnet mask as
described in “Setting the YH8000 IP address” on the next page before reconnecting.
(1) Tap
in the lower right of the screen.
The tag names (or the serial numbers if not assigned) of the connected TDLS8000s appear.
Select the TDLS8000 that you want to change the IP address of.
In the example in the figure, the tag name is “TAG1.” Tag names are not assigned to
TDLS8000s with factory default settings, so serial numbers are displayed instead.
After selecting, tap OK. A password input screen will appear.
(2) Enter the password and tap Enter. A configuration screen will appear. The factory default
password is “1234.”
From the menu, select Configuration>>System>>Communication>>TCP/IP. A warning
screen explaining that the instrument will restart if the IP address is changed appears.*1 Tap
OK.
*1:
The analyzer will restart after changing IP settings”
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
5-6
<5. Startup>
(3) Select each item and change the value. The items that you change will show an asterisk in
the upper left. Tapping OK will display a confirmation screen for restarting. Tap OK again.
(4) A disconnection error screen will appear. Tap OK.
(5) The YH8000 (HMI) configuration screen will appear. Tap Analyzer Connection.
(6) The TDLS8000 selection screen will appear. Tap Disconnect.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
5-7
<5. Startup>
(7) The Change IP button becomes available. Tap Change IP, and enter the new destination IP
address.
(8) Tap Connect to connect to the TDLS8000.
NOTE
The TDLS8000 IP address can also be changed via HART communication.
Configuration menu path:
“Detailed setup>>System>>Communication>>TCP/IP>>Set IP settings”
l Setting the YH8000 IP address
Configuration menu path: “
>>HMI>>Network Setting”
NOTE
When you change the YH8000 IP address, the YH8000 automatically restarts with the new IP
address, disconnecting the current connection with the TDLS8000.
5.2.3
Connecting to the TDLS8000
This section explains the TDLS8000 selection screen on the YH8000. From the YH8000 screen,
you can connect in the following ways.
• Connect to a TDLS8000 with any IP address
• Connect up to four TDLS8000s from a single YH8000
• Connect to a TDLS8000 with an unknown IP address in the same network by searching
The YH8000 assigns an analyzer number from 1 to 4 to each connected TDLS8000. In the case
of a one-to-one connection, analyzer number 1 is used. On the TDLS8000 selection screen, you
can assign TDLS8000s to analyzer numbers by specifying the IP address.
Path to the TDLS8000 selection screen “
>>HMI>>Analyzer Connection”
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
5-8
<5. Startup>
The procedure to add a connection destination is shown below.
(1) Open the TDLS8000 selection screen (the figure below), and select the analyzer number
that you want to set under the Analyzer No. column.
(2) Tap Change IP to open the IP address configuration screen (the figure below).
If the selected analyzer number is already connected (Status is Connected), tap Disconnect
to enable the Change IP button.
(3) Specify the TDLS8000 IP address using either method below.
• Input Analyzer IP manually
Manually enter the IP address of the TDLS8000 you want to connect to.
• Select Analyzer by Auto-search
Select from the TDLS8000 IP addresses automatically found. Up to 32 TDLS8000s in
the same subnet can be detected. Detection may not be possible depending on the
network configuration or condition.
Select either option, tap OK, and specify the IP address you want to connect to.
(4) Tap Connect to connect to the TDLS8000.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
5-9
<5. Startup>
l Description of the TDLS8000 selection screen
Analyzer No. column
Shows analyzer numbers 1 to 4. You can select a number and assign an IP address to
connect to the corresponding TDLS8000.
NOTE
If there is only a single TDLS8000 that the YH8000 will connect to, connect it to analyzer number
1. Otherwise, you will need to reset the display items when displaying the trend waveform.
Analyzer Information column
Shows the TDLS8000 IP addresses and tag names. For TDLS8000s that have not been
connected before, tag names do not appear. If a TDLS8000 is connected but a tag name is
not assigned, its serial number will be displayed.
Status column
Shows the TDLS8000 connection status.
Connected
Connecting
Disconnected
Change IP button
Switches to the IP address configuration screen. This button is not available if the
TDLS8000 is connected.
Connect button
Starts a connection with the TDLS8000. This button is available when the TDLS8000 is
disconnected.
Disconnect button
Disconnects from the TDLS8000 that is connected or is in the process of establishing a
connection.
NOTE
Once a connection is established between the YH8000 and TDLS8000, the connection
information is saved. Therefore, the next time the power is turned on or after restarting, the
connection will be established automatically.
NOTE
Up to two YH8000s can connect to a single TDLS8000.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
5-10
<5. Startup>
5.2.4
Handling Connection Failures
If a connection to the TDLS8000 fails, check the following items.
(1) Check that the Ethernet cable is connected properly.
LED5
• Check that LED5 on the SCU circuit board is lit or blinking.
Figure 5.2
Front view of the inside of the SCU
• Check that the green LED of the YH8000 Ethernet port is lit or blinking.
Green LED
Figure 5.3
Top view of the inside of the YH8000
(2) Check that the IP addresses are set properly.
5.2.5
•
Check that unique IP addresses are assigned to each YH8000 and TDLS8000.
=> You can view the TDLS8000 IP addresses on the SCU display.
For details, see “1.2 Name and Function of Each Part”.
•
Check that the subnet addresses and subnet masks are set properly.
•
If the instruments are connected via a router or the like, check that it is configured
properly with your network administrator.
Basic Screen Configuration
This section provides an overview of the home screen that appears when a connection is
established with a TDLS8000 and the buttons. For a detailed explanation, see “8.2
Home
Screen”.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
5-11
<5. Startup>
The screen shown in Figure 5.4 appears when a single TDLS8000 is connected to the YH8000.
This screen is called the home screen.
Date and time Operating status
Tag name
HART communication status
Concentration
Concentration meter
Transmission meter
Transmission
Process pressure
Process temperature
Pressure icon
Temperature icon
Buttons at the bottom of the screen
Figure 5.4
Home screen
If several TDLS8000s are connected to the YH8000, tabs for switching between the home
screens of each TDLS8000 appears on the right. The tab numbers represent analyzer numbers.
Selecting a tab shows the information of the TDLS8000 corresponding to the tab number.
Further, the tab shows the information of all connected TDLS8000s. This screen is called the
overall screen.
Date and time
Display switch tab
Analyzer number
Temperature icon
Pressure icon
Alarm
Icon
Figure 5.5
Overall screen
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
5-12
<5. Startup>
l Functions of the buttons at the bottom of the screen
These buttons are used to show other screens.
Button
Button name
Description
Home button
Displays the home screen.
Trend display
button
Alarm
information
button
Displays the trend screen. You can view the measurement
trends of multiple TDLS8000s on the same screen.
Configuration
display button
Configuration
button
Displays the TDLS8000 alarm screen.
Displays the current TDLS8000 settings. The button can
also be used to display I/O values, alarm history, and so
on.
Executes TDLS8000 configuration, calibration, and the
like. It can also be used to change the YH8000 connection
destination, IP settings, and so on.
l How to open the TDLS8000 configuration screen
(1) Tap
to show a screen for selecting the TDLS8000 you want to configure. TDLS8000 tag
names (serial numbers if tag names are not assigned) are displayed.
Select the target TDLS8000, and tap OK.
(2) A user password input screen will appear. The factory default password is “1234.”
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<5. Startup>
5-13
(3) When password authentication is successful, a TDLS8000 configuration screen will appear.
NOTE
If you just want to view the settings, you can tap
need to enter a password to display this screen.
5.3
to display the information screen. You don’t
Setting Basic Parameters
This section explains how to set the basic parameters necessary to start measurements. Field
Communicator 475 will be used as an example of a HART configuration tool to introduce the
operation procedure. If you are using FieldMate or other tool, perform the equivalent procedure.
NOTE
When configuration changes are made with a Field Communicator 475, the following screen will
appear. This warning screen is to indicate that the TDLS8000 settings have been changed. Since
this is not a problem with the TDLS8000 operation, select OK and then YES.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
5.3.1
5-14
<5. Startup>
Setting the Date and Time
Set the current date and time on the TDLS8000. The date and time will be retained through
battery power even when the power is turned off.
l Setup procedure using HART
(1) Select “Detailed setup>>System>>Date/time>>Set date/time” and press the right arrow
button to display the following screen.
(2) The current time will appear. If you want to change it, tap OK. The following screen will
appear. If you do not need to change it, tap ABORT to abort the setup.
Enter the current date in yyyy/mm/dd format, and then tap ENTER.
(3) Enter the current time in hh:mm:ss format, and then tap ENTER. To enter a colon, tap the
@& key.
Note: The time you enter here is reflected in the TDLS8000 when you tap OK in step (4).
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<5. Startup>
5-15
(4) A screen appears for you to confirm the date and time that will be applied to the TDLS8000.
To proceed, tap OK. The time will be reflected in the TDLS8000. To abort the setup, tap
ABORT.
l Setup procedure using YH8000
(1) Tap
to switch to the TDLS8000 selection screen. Select the tag name of the TDLS8000
you want to connect to, and then tap OK.
(2) A password input screen will appear. Enter the password and tap Enter. A configuration
screen will appear.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
5-16
<5. Startup>
(3) Select “Configuration>>System>>Date&Time” as shown in the following figure.
(4) Select “Date” and “Time,” and enter the date and time you want to apply.
(5) The items that you change will show an asterisk in the upper left as shown in the following
figure. Tap OK to reflect the date and time in the TDLS8000.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<5. Startup>
5.3.2
5-17
Setting the Process Optical Path Length
Set the process optical path length. The process optical path length is the length that the laser
beam travels through the gas to be measured. As shown in the following figure, enter the length
(L) of the process that the LU and the SCU are connected to. It does not include the length of the
flanges.
If insertion tubes are installed, the optical path length is the length between the closer ends of the
insertion tubes.
Process optical path length (L)
Insertion tubes
LU
SCU
Process optical path length when insertion tubes are installed
Figure 5.6
Process optical path length definition
If the LU and SCU are mounted on a free cell, enter the length of the flow cell as the process
optical path length.
l Setup procedure using HART
(1) From the menu, select “Detailed setup>>Analysis>>Process param.” The following screen
will appear. Select “OPL” and press the right arrow button.
(2) Enter the process optical path length, and tap ENTER to execute writing to the TDLS8000.
To abort, tap ESC.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
5-18
<5. Startup>
l Setup procedure using YH8000
Configuration menu path:
“
>>Configuration>>Analysis>>Process Parameters>>Path Length”
NOTE
If you are not using nitrogen purge gas, you also need to enter the optical path length, pressure,
and temperature of the non-measured process. For details, see “6.10 Non-process Parameter
Settings”.
5.3.3
Setting the Process Pressure
This section explains the pressure value of the measurement process when the input mode is
analog input (AI-1). When using other input modes, see “6.1.2 Process Pressure”.
Configuration menu path:
[HART] “Detailed setup>>Analysis>>Process param>>Pressure”
[YH8000] “
>>Configuration>>Analysis>>Process Parameters>>Pressure”
(1) Set Mode (pressure input mode) to Active input.
(2) Set Active type (pressure input source) to AI-1.
(3) Select Backup mode (pressure value backup when the analog input is outside the range)
from the following options.
If you select Disable, the analog input is converted as-is into a pressure value without
backing up.
If you select Backup value, the pressure value is fixed to the value specified by Backup set
value. Set Backup set value to a value of your choice.
If you select Hold, the pressure value is held at the previous value within the proper range.
(4) Set the analog input range. Set the pressure values corresponding to 4 mA and 20 mA.
Configuration menu path:
[HART] “Detailed setup>>I/O condition>>Analog input>>AI-1(Pressure)”
[YH8000] “
5.3.4
>>Configuration>>I/O>>Analog Input>>AI-1(Pressure)”
Setting the Process Temperature
This section explains the temperature value of the measurement process when the input mode is
analog input (AI-2). When using other input modes, see “6.1.3 Process Temperature”.
Configuration menu path:
[HART] “Detailed setup>>Analysis>>Process param>>Temperature”
[YH8000] “
>>Configuration>>Analysis>>Process Parameters>>Temperature”
(1) Set Mode (temperature input mode) to Active input.
(2) Set Active type (temperature input source) to AI-2.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
5-19
<5. Startup>
(3) Select Backup mode (temperature value backup when the analog input is outside the
range) from the following options.
If you select Disable, the analog input is converted as-is into a temperature value without
backing up.
If you select Backup value, the pressure value is fixed to the value specified by Backup set
value. Set Backup set value to a value of your choice.
If you select Hold, the pressure value is held at the previous value within the proper range.
(4) Set the analog input range. Set the temperature values corresponding to 4 mA and 20 mA.
Configuration menu path:
[HART] “Detailed setup>>I/O condition>>Analog input>>AI-2(Temperature)”
[YH8000] “
5.3.5
>>Configuration>>I/O>>Analog Input>>AI-2(Temperature)”
Setting the Output Range
This section explains how to assign an item to the 4 to 20 mA analog output. For details on the
analog output hold function, see “6.4.2 Output Hold”.
Configuration menu path:
[HART] “Detailed setup>>I/O condition>>Analog output”
[YH8000] “
>>Configuration>>I/O>>Analog output”
(1) Select the output item for each channel.
On HART, the AO-1 and AO-2 output items are displayed as “PV is” and “SV is,”
respectively. On YH8000, it is displayed as “Item.” You can assign the following items to the
analog output.
Output item
Component 1 gas
concentration
Component 2 gas
concentration (*1)
Transmission
Temperature
Pressure
*1:
*2:
Name displayed on Name displayed on
HART
YH8000
Concentration 1
(*2)
Concentration 2
(*2)
Transmission
Temperature
Pressure
Transmission
Temperature
Pressure
Selectable only with two-gas measurement specifications
Displays the name of the gas component being measured.
Example: If the component being measured is oxygen: O2 concentration
(2) Set the measurement item values that correspond to the minimum point (4 mA) and
maximum point (20 mA).
On HART, the minimum and maximum points are displayed as “PV LRV” and “PV URV,”
respectively.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
5-20
<5. Startup>
5.3.6
Setting Process Alarms
You can set threshold values of the high/low limit alarm (warning) for process measurement
values. The following table shows the types of warnings that you can specify. For each type of
warning, you can set the threshold values and select whether to enable the detection. For details
on warnings, see “10.2 Warning Display and Handling” described later.
Alarm number
1
2
3
4
5
6
7
8
9
Warning name
Transmission Low
Process Pressure Low
Process Pressure High
Process Temperature Low
Process Temperature High
Concentration Gas1 Low
Concentration Gas1 High
Concentration Gas2 Low (only with two-gas measurement)
Concentration Gas2 High (only with two-gas measurement)
l Setup procedure using HART
(1) From the menu, select “Detailed setup>>Alarm>>Warning>>Warning group 1.”
The following screen will appear.
(2) “Warn select group 1” is used to enable or disable the warnings. By factor default, all
warnings are enabled.
Select “Warn select group 1” and press the right arrow button to display the following
screen. Here, you can enable or disable each warning.
ON means enabled. To switch between ON and OFF, tap OFF (center-left at the bottom of
the screen). When you are finished, tap ENTER.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
5-21
<5. Startup>
(3) Set the high and low limit threshold values of each warning.
For example, to change the lower limit alarm threshold value for transmission, select “Trans
low lmt,” and press the right arrow button. The following screen will appear.
Enter the value, and press ENTER.
(4) Tap SEND to execute writing to the TDLS8000.
(5) For two-gas measurement, select “Detailed setup>>Alarm>>Warning>>Warning group
1,” and set the gas concentration low limit alarm for component 2. Then select “Detailed
setup>>Alarm>>Warning>>Warning group 2,” and set the gas concentration high limit
alarm.
l Setup procedure using YH8000
(1) From the menu, select “
>>Configuration>>Alarm” to display the following screen.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<5. Startup>
5-22
(2) You can enable or disable each warning by tapping the check mark on the left side. In the
following example, the “Transmission Low” check box is selected to enable the transmission
low limit alarm. Touch Apply to apply the settings to the TDLS8000.
(3) Change the threshold values of each warning. In the following example, to set the threshold
value of the transmission low limit alarm, select “Transmission Low,” and tap Setting on the
right side.
(4) The following screen will appear. Select the value box, enter the threshold value, and touch
OK to apply the value to the TDLS8000.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
5.4
5-23
<5. Startup>
Loop Check (Simulation output)
You can force the analog output, digital output, and valve control output to a given state. This
section explains how to do this. You can use this function to check the operation after wiring.
5.4.1
Executing a Loop Check
This section explains the loop check setup procedure for each output type separately. A loop
check can be executed simultaneously on all terminals of all types.
Execution menu path:
[HART] “Diagnosis/Service>>Loop check”
[YH8000] “
>>Execution>>Loop Check”
NOTE
If you turn off the TDLS8000 while performing a loop check, loop check will be cleared.
l Analog output
Open the Analog output menu, and set Loop check mode to Enable to output the specified
simulated current (“check output”).
NOTE
If multi-drop mode is set on HART, loop checking of AO-1 via HART is not possible.
l Digital output
Open the Digital output menu, and set Loop check mode to Enable to output the specified
simulated state (“check output”).
l Valve control output (SV)
Open the Valve output menu, and set Loop check mode to Enable to output the specified
simulated state (“check output”).
5.4.2
Auto Release Function
The auto release function automatically clears loop checking on all terminals and restores normal
output after the specified time elapses. The auto release counter starts when any of the loop
check is enabled. The counter restarts whenever a loop check setting is changed. When the
counter expires, all loop checking is disabled.
To set the auto release time, set Auto release time in the Loop check menu. If you select Disable,
the auto release function is disabled, and the simulated output is maintained until you manually
clear the loop check mode. The specified values are retained even after the power is turned off.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
Blank Page
6.
<6. Configuration>
6-1
Configuration
This chapter provides details of all the setting items and shows the locations of the setting menus
of the TDLS8000. However, the setting items related to calibration and validation are described in
“9.
Inspection and Maintenance”.
The setting menus of the YH8000 can be opened by performing the operation of “
>> Select
analyzer >>Configuration” but the location is shown as just being under “Configuration” in this
chapter for simplicity.
6.1
Process Parameter Settings
Process parameters indicate the measurement conditions related to measurement process gas.
Set the process optical path length, process pressure, and process temperature of the process
gas correctly because they directly affect the measurement values.
NOTE
If nitrogen purge gas will not be used, you also need to configure the non-measurement process
optical path length, pressure, and temperature in addition to the settings in this section. For
details, see “6.10
Non-process Parameter Settings”.
6.1.1
Process Optical Path Length
Set the process optical path length of the process to be measured. For a definition of the process
optical path length, see “5.3.2 Setting the Process Optical Path Length”.
Setup menu path:
[HART] “Detailed setup>>Analysis>>Process param>>OPL”
[YH8000] “Configuration>>Analysis>>Process Parameters>>Path Length”
6.1.2
Process Pressure
Set the process pressure of the process to be measured. When you select the input source, set
the action for error input as described in the procedure below.
Setup menu path:
[HART] “Detailed setup>>Analysis>>Process param>>Pressure”
[YH8000] “Configuration>>Analysis>>Process Parameters>>Pressure”
(1) Select “Mode” (pressure input mode).
When “Fixed” is selected, set the pressure value as an arbitrary fixed value.
When “Active input” is selected, input the pressure value via a 4-20 mA analog input (AI-1)
or Modbus. The TDLS8000 acquires a pressure value sent from the input source at every
measurement value analysis cycle and uses it for the concentration calculation.
(2) Configure the settings below according to the pressure input mode.
When “Fixed” is selected
Enter a fixed pressure value for “Fixed mode value.” The other parameters do not need to
be set.
When “Active input” is selected
Set each of the following parameters.
• Active type: Setting of pressure input source
“AI-1” is analog input (AI-1).
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
6-2
<6. Configuration>
“Field Communication” is Modbus input. For details on the Modbus function, see “11.
Modbus”.
• Backup mode: When “Active input” is selected, set the pressure value to use for the
concentration calculation when an input value from the sensor is out of range.
With “Disable,” backup is disabled and the input is used as is for the pressure value.
With “Backup value,” the value set in “Backup set value” is used for the pressure
value.
With “Hold,” the intermediate value of the input values of five cycles that were within
the range immediately before an input value from the sensor became out of range
is used for the pressure value. The following figure shows the example of when the
analog input falls below the lower limit of 4 mA.
A ● in the figure indicates the point of acquisition of an AI input value, and a value is
acquired every analysis cycle. The second ● of the five cycles immediately before
falling below the lower limit is the intermediate value of the current values so the
pressure value of this time is held. If the analog input immediately after turning on
the power of the TDLS8000 is out of range while “Hold” is selected, a pressure value
corresponding to 4 mA is held.
AI input current [ mA ]
Intermediate value
4.0
Time [ s]
Analysis cycle
• Backup set value: Backup pressure value when “Backup mode” is “Backup value”
(3) Configure the analog input range setting only when “AI-1” is selected in “Active type.”
Enter the pressure value corresponding to each of 4 mA and 20 mA.
Setup menu path:
[HART] “Detailed setup>>I/O condition>>Analog input>>AI-1(Pressure)”
[YH8000] “Configuration>>I/O>>Analog Input>>AI-1 (Pressure)”
6.1.3
Process Temperature
Set the process temperature of the process to be measured. When you select the input source,
set the action for error input as described in the procedure below.
Setup menu path:
[HART] “Detailed setup>>Analysis>>Process param>>Temperature”
[YH8000] “Configuration>>Analysis>>Process Parameters>>Temperature”
(1) Select “Mode” (temperature input mode).
When “Fixed” is selected, set the temperature value as an arbitrary fixed value.
When “Active input” is selected, input the temperature value via a 4-20 mA analog input (AI2) or Modbus. The TDLS8000 acquires a temperature value sent from the input source at
every measurement value analysis cycle and uses it for the concentration calculation.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
6-3
<6. Configuration>
When “Active ambient” is selected, the value of the temperature sensor mounted in the
vicinity of the laser device in the TDLS8000 is used as the process temperature.
(2) Configure each of the settings below according to the temperature input mode.
When “Fixed” is selected
Enter a fixed temperature value for “Fixed mode value.” The other parameters do not need
to be set.
When “Active input” is selected
The setting items are the same as those in section 6.1.2. Read “pressure” as “temperature”
and “AI-1” as “AI-2” when configuring the settings.
When “Active ambient” is selected
Set the temperature offset value in “Temp act amb ofst” (“Offset value” on the YH8000).
The temperature sensor value is the temperature in the vicinity of the laser device, and is
not exactly equal to the process temperature so set a difference as an offset.
(3) Configure the analog input range setting only when “AI-2” is selected in “Active type.”
Enter the temperature value corresponding to each of 4 mA and 20 mA.
Setup menu path:
[HART] “Detailed setup>>I/O condition>>Analog input>>AI-2(Temperature)”
[YH8000] “Configuration>>I/O>>Analog Input>>AI-2 (Temperature)”
6.2
Unit Settings
Set the units for physical quantities related to concentration measurement. The physical
quantities for which units can be set are “optical path length,” “pressure,” and “temperature,” and
you can select from the following units for each of them.
When Only SI unit (-J) is assigned to TDLS8000, non SI Unit will not be displayed.
Item
SI
Optical path length m, cm
Pressure
kPa, atm, bar
inch, feet
psi, torr
Temperature
deg F
deg C, K
Setup menu path:
[HART] “Detailed setup>>Analysis>>Unit”
[YH8000] “Configuration>>Analysis>>Units”
6.3
Analog Input Settings
Analog input is used for the purpose of calculating a pressure value and temperature value from
analog input within the 4-20 mA range. Set the pressure value and temperature value ranges
here which will correspond to 4 mA and 20 mA.
Setup menu path to A1-1 (pressure input):
[HART] “Detailed setup>>I/O condition>>Analog input>>AI-1(Pressure)”
[YH8000] “Configuration>>I/O>>Analog Input>>AI-1(Pressure)”
Setup menu path to A1-2 (temperature input):
[HART] “Detailed setup>>I/O condition>>Analog input>>AI-2(Temperature)”
[YH8000] “Configuration>>I/O>>Analog Input>>AI-2(Temperature)”
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
6-4
<6. Configuration>
NOTE
For how to set the pressure values and temperature values for analog input, see “6.1.2 Process
Pressure” and “6.1.3 Process Temperature”.
6.4
Analog Output Settings
This section describes how to set the process measurement values for analog output and the
function to hold output in accordance with the status of the TDLS8000.
6.4.1
Normal Range Output
This section describes how to set 4-20 mA analog output and the detailed operation.
Setup menu path:
[HART] “Detailed setup>>I/O condition>>Analog output>>AO-1 or AO-2”
[YH8000] “Configuration>>I/O>>Analog Output>>AO-1 or AO-2>>Output”
l Output items and range settings
(1) Select the measurement item to assign to analog output.
Output Item
1st component gas concentration
2nd component gas concentration (*1)
Transmission
Temperature
Pressure
*1:
*2:
HART Display Name
Concentration 1
Concentration 2
Transmission
Temperature
Pressure
YH8000 Display Name
(*2)
(*2)
Transmission
Temperature
Pressure
Only selectable for two-gas measurement specification
Displays the gas name of the measurement component
(Example) When the measurement component is oxygen: O2 Concentration
(2) Enter the values for the measurement items corresponding to the lower range value (4 mA)
and upper range value (20 mA). In the case of HART, each of the lower range value (LVR)
and upper range value (URV) is displayed.
l Output value at startup
The analog output value is fixed to 4.0 mA during the period from after turning on the power of
the TDLS8000 until the first measurement result is updated. However, if output hold is set for the
warming-up state, the value is in accordance with that setting.
l Analog output range
Measurement values within the range from 3.8 mA to 20.5 mA are output (NAMUR NE43
compliant).
6.4.2
Output Hold
Output hold is a function to fix (hold) analog output to a set value when the TDLS8000 is in the
following specific states.
Setup menu path:
[HART] “Detailed setup>>I/O condition>>Analog output>>AO-1 or AO-2>>Hold menu for
each specific state”
[YH8000] “Configuration>>I/O>>Analog Output>>AO-1 or AO-2>>Hold Mode”
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
6-5
<6. Configuration>
l Definitions of specific states
Set output hold individually for each of the following states.
During fault occurrence
State when any fault is occurring
During warning
State when any warning is occurring
occurrence
During calibration and
State when either the calibration or validation function is being executed
validation
During maintenance
State in which the password for maintenance has been entered from the
YH8000 and changing of the settings is enabled
During warm-up
State up until the temperature of the laser device stabilizes and
measurement becomes possible after turning on the power of the
TDLS8000.
l Output hold mode
One of the following modes can be selected for output hold.
Setting Mode
Preset hold
Description
Holds output to any output value within 3.8 to 20.5 mA.
* Any value within 3.0 to 21.6 mA can be set for during warning occurrence
and during fault occurrence. Furthermore, the holding of output to any
value can be delayed for a period of up to five analysis cycles (*1).
Analog output during this delay is held to the value immediately before
just as with the Hold mode.
Analog output is not held. Measurement values continue to be output.
Holds output to the normal output value immediately before.
Non-hold
Hold
*1:
The analysis cycle differs depending on the type of analyzer used. For details on the analysis cycle, see “Appendix 5 What Is an
Analysis Period?”.
(Example 1) Figure showing the analog output action when Preset hold is set to 21.0 mA and
the number of delays is set to 2 for during fault occurrence
AO current [ mA ]
21
Fault
occurrence
0
Time [ s]
Analysis cycle
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
6-6
<6. Configuration>
(Example 2) Figure showing the analog output action when Preset hold is set to 21.0 mA and
the number of delays is set to 0 for during fault occurrence
AO current [ mA ]
21
Fault
occurrence
0
Time [ s]
Analysis cycle
l Output hold priority
When multiple specific states occur at the same time and multiple holds are enabled (when
multiple Preset holds or Hold modes are enabled), the output hold is determined according to the
following priority order.
Priority High
During fault occurrence
During warning occurrence
During calibration and validation
During maintenance
During warm-up
(Example) When during warning occurrence = Non-hold, during maintenance = Preset
hold, and during calibration and validation execution = Hold, and all three of the
aforementioned states occurred at the same time, output hold becomes the Hold
mode.
6.5
Digital Output Settings
This function is for turning on digital output when the TDLS8000 enters the following specific
states.
6.5.1
DO Contact (DO-1)
Setup menu path:
[HART] “Detailed setup>>I/O condition>>Digital output>>DO-1(DO)”
[YH8000] “Configuration>>I/O>>Digital Output>>DO-1(DO)”
l Definitions of specific states
When any of the following specific states occurs, the contact turns on. Enabling or disabling of
digital output can be set separately for each specific state. For an explanation of each state, see
“6.4.2 Output Hold”.
• During warning occurrence
• Calibration and validation
• During maintenance
• During warm-up
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
6-7
<6. Configuration>
l Output delay
A delay of up to 100 analysis cycles can be set for the period from when a specific state occurs
until when the contact actually turns on. When the number of delays is set to zero, the contact
turns on immediately after a state occurs.
The analysis cycle differs depending on the application and is set to the optimum value at the
time of shipment. For details, see “Appendix 5 What Is an Analysis Period?”.
6.5.2
Fault Contact (DO-2)
Setup menu path:
[HART] “Detailed setup>>I/O condition>>Digital output>>DO-2(Fault)”
[YH8000] “Configuration>>I/O>>Digital Output>>DO-2(Fault)”
l Definitions of specific states
When a fault occurs, the contact turns on. This contact is specifically for fault notification and
cannot be disabled.
l Output delay
A delay of up to 100 analysis cycles can be set for the period from when a fault occurs until when
the contact actually turns on. When the number of delays is set to zero, the contact turns on
immediately after a fault occurs.
The analysis cycle differs depending on the application and is set to the optimum value at the
time of shipment. For details, see “Appendix 5 What Is an Analysis Period?”.
6.6
Process Alarm Settings
Of the alarms of the TDLS8000, the threshold value and enable and disable can be set arbitrarily
only for the following warnings related to the measurement process status. For details on each
alarm, see “10.2
Warning Display and Handling”.
Alarm No.
1
2
3
4
5
6
7
8
9
Alarm Name
Transmission low
Pressure low
Pressure high
Temperature low
Temperature high
Concentration gas1 low
Concentration gas1 high
Concentration gas2 low (analyzer 2 only)
Concentration gas2 high (analyzer 2 only)
Setup menu path:
[HART] “Detailed setup>>Alarm”
[YH8000] “Configuration>>Alarm”
For details on the setting procedure, see “5.3.6 Setting Process Alarms”.
6.7
Digital Input Settings
A specific function can be executed depending on the digital input (DI-1, DI-2). Also, the function
to execute can be set for each channel. There are two types of digital input methods, edge input
and status input, and the input method differs depending on the function.
• Edge input
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
6-8
<6. Configuration>
Execute a function when the digital input changes from “open” to “closed.”
• Status input
Execute and continue executing a function while the digital input is closed.
Setup menu path:
[HART] “Detailed setup>>I/O condition>>Digital input>>DI-1 or DI-2”
[YH8000] “Configuration>>I/O>>Digital Input>>DI-1 or DI-2”
l Functions that can be executed
The following shows the functions that can be executed.
Function Name
External Alarm
Zero Calibration
Span Calibration
Zero + Span
Calibration
Offline Validation 1
Offline Validation 2
Offline Validation 1 + 2
Online Validation 1
Online Validation 2
Stream 1
Stream 2
Stream 3
Action
Generates the “External Alarm” warning.
Executes automatic zero calibration.
Executes automatic span calibration.
Continuously executes automatic zero calibration and
automatic span calibration.
Executes automatic offline validation 1.
Executes automatic offline validation 2.
Continuously executes automatic offline validation 1 and 2.
Executes automatic online validation 1.
Executes automatic online validation 2.
Switches the valve to stream 1.
Switches the valve to stream 2.
Switches the valve to stream 3.
Input Method
Status
Edge
Edge
Edge
Edge
Edge
Edge
Edge
Edge
Status
Status
Status
Note: The items of the above selection options that have been disabled by the “Valve usage setting” are not displayed. For details, see
“6.8.2
Valve Usage Setting”.
l Filter time
A cutoff time can be set to prevent wrong operation due to chattering. A digital input change within
the specified time will be ignored.
6.8
Valve Stream Settings
This section describes the procedure to automatically control multiple process gas streams
according to the TDLS8000 valve control output (SV terminal). Up to three streams can be
switched.
6.8.1
Definitions of Stream Numbers
Implement the valve operation of the TDLS8000 using stream numbers defined for the statuses
of the two valves connected to the valve control outputs (SV-1 and SV-2). Do not perform valve
operation independently but specify stream numbers (independent operation is possible only
when performing a loop check).
The following shows the stream numbers to define for the TDLS8000 and the corresponding
status of each valve.
Stream No. Valve 1 (SV-1) Status Valve 2 (SV-2) Status
Stream 1
OFF
OFF
Stream 2
ON
OFF
Stream 3
ON
ON
The process gases can be switched as shown in Figure 6.1 by switching the streams.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<6. Configuration>
Laser unit (LU)
6-9
Sensor control unit (SCU)
Flow cell
Purge gas
Purge gas
Process gas 1
(Stream1)
Valve-2
(SV-2)
Valve-1
(SV-1)
Process gas 3
(Stream3)
Figure 6.1
6.8.2
Process gas 2
(Stream2)
Piping diagram for switching streams
Valve Usage Setting
Set the valve usage purpose for automatically controlling the valves according to the SV terminal.
“3 Streams Switching,” “2 Streams & Cal Val),” and “Cal/Val” can be selected.
If “3 Streams Switching” is selected, three process gases can be switched and measured in
order. In the case of 2 stream switching, a calibration gas or check gas can be connected to the
remaining stream (stream 3) so that it can be used for calibration or validation.
Setup menu path to Valve usage:
[HART] “Detailed setup>>I/O condition>>Valve control>>Set valve usage”
[YH8000] “Configuration>>I/O>>Valve Control>>Valve Usage”
NOTE
The stream switching and automatic calibration and validation items that can be executed
according to the TDLS8000 valve control output (SV terminal) are limited depending on the
“Valve usage” setting.
• Cal/Val Usage
Stream switching cannot be used because all streams are used for automatic calibration
and validation. When calibration and validation are not executed, the stream is fixed to
stream 1.
To switch the process gas, do it manually.
• 2 Streams & Cal/Val Usage
Sets only stream 3 to be used for automatic calibration and validation. The following
automatic calibration and validation cannot be executed because they would use stream 2.
Calibration and validation can only be executed manually.
• Zero calibration
• Offline validation 1
• Online validation 1
• 3 Streams Switching Usage
All streams are set to be used for stream switching. None of the automatic calibration or
validation can be executed. Calibration and validation can only be executed manually.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
6-10
<6. Configuration>
NOTE
If valve usage is changed, the following setting values are initialized.
• The time execution setting (Time initiate) of all automatic calibration and validation is set to
“Disable.”
• If remote execution of automatic calibration and validation is set for digital input, it is set to
“Disable.”
• If stream switching is set for digital input, it is set to “Disable.”
• The “Current stream” setting is set to “Stream 1.”
• The “Initial stream” setting is set to “Stream 1.”
6.8.3
Stream Settings
There are the following three stream switching methods.
Manual
Time elapse
Digital input
Specify a stream and then switch to it from an YH8000 or HART screen.
Switches to the next stream automatically when the specified time elapses after a stream has
been switched. When switching is performed manually or when switching to the initial stream
is performed after the power of the TDLS8000 is turned on, counting of the specified time
begins. This function does not work for switching by digital input.
Switches to the specified stream while the digital input is closed. The specified stream is
switched to giving priority over the stream switched to with the manual or time elapse method.
When the digital input returns to the open state, the stream returns to that switched to with the
manual or time elapse method. The following figure shows the priority order.
The following figure shows the priority order. The manual and time elapse methods have the
same priority level, and operation of the digital input is given higher priority.
Highest priority level
Digital input
Manual
Time elapse
l Switching by manual operation
Setup menu path:
[HART] “Detailed setup>>I/O condition>>Valve control>>Current stream”
[YH8000] “Configuration>>I/O>>Valve Control>>Current”
To switch the stream manually, select the stream you wish to switch to in “Current stream.”
l Switching by time elapse
If “3 Streams Switching” or “2 Streams & Cal Val” is selected for the “Valve usage” setting, the
valve open duration and the stream to switch to can be set for each stream. Enter the duration to
open the valve (duration for gas to flow to the target stream) in “Duration” and specify the stream
you wish to switch to next in “Switch to.” Setting the next stream in “Switch to” in the same way
for the stream that is switched to enables switching again to that stream after the specified time
elapses.
Setup menu path:
[HART] “Detailed setup>>I/O condition>>Valve control>>Stream time switch”
[YH8000] “Configuration>>I/O>>Valve Control”
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<6. Configuration>
6-11
[Example of stream switching]
When “Valve usage” = “3 Streams Switching” and the settings are as follows:
Stream
Current Switch to Duration
1
2
3
■
□
□
Stream 2
Stream 3
Disable
Initial
10
20
30
■
□
□
The transition to the current stream will be as shown in the figure below.
10 minutes
Power supply
1
Current stream
2
3
Duration
l Switching by digital input
Set the function to switch the stream upon digital input. For the setting procedure, see “6.7 Digital
Input Settings”.
[Example of switching by digital input]
When “Valve usage” = “3 Streams Switching,” “DI-1 Mode” = “Stream 3,” and the settings are as follows:
Stream
1
2
3
Current Switch to Duration
■
□
□
Stream 2
Stream 3
Disable
Initial
10
10
10
■
□
□
“Current stream number” in the following figure refers to the final stream. Priority is given to
switching by digital input so the stream is switched to stream 3 while DI-1 is closed. However,
since counting and stream switching by the time elapse method continues to be performed
internally (bottommost line in the figure), the stream switches to stream 2 at the point when DI-1
returns to the closed state.
10 minutes
Power supply
DI-1
Current stream number
Duration
Switching by
time elapse
6.8.4
1
3
2
3
Switching by digital input
1
2
3
Initial Stream (Stream at Startup)
When the power of the TDLS8000 is turned on, the current stream is switched to the stream set
as the initial stream. The settable range of the “Initial stream” setting is only a stream selected
for stream switching usage in the “Valve usage” setting. The streams that can be set as the initial
stream are indicated by ○ in the table below.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<6. Configuration>
Valve usage
Cal/Val usage
2 Streams & Cal/Val usage
3 Streams Switching usage
6-12
Stream 1 Stream 2 Stream 3
○
○
○
×
○
○
×
×
○
Setup menu path:
[HART] “Detailed setup>>I/O condition>>Valve control>>Initial stream”
[YH8000] “Configuration>>I/O>>Valve Control>>Initial”
NOTE
If switching the stream by time elapse is set for the stream number of the initial stream, counting
of the switching time begins from after the power is turned on.
6.9
Other Settings
This section describes various settings other than measurement process settings and the I/O and
alarm settings.
6.9.1 Tag
This is a tag of up to 32 ASCII characters for identifying individual TDLS8000. It is displayed when
you connect to the TDLS8000 from the YH8000. Furthermore, the long tag defined as standard
in HART communication is the same as this tag. From YH8000, Latin-1 character cannot be
inputted on tag.
Setup menu path:
[HART] “Detailed setup>>System>>Long tag”
[YH8000] “Configuration>>System>>Tag”
6.9.2
Date and Time
Set the current date and time. For a detailed description of the setting screen, see “5.3.1
Setting the Date and Time”. Furthermore, it is possible to set the time simultaneously for multiple
connected TDLS8000 only when setting it from the YH8000. For details, see “8.7.3
Setting
the Date and Time on the TDLS8000”.
Setup menu path:
[HART] “Detailed setup>>System>>Date/time>>Set date/time”
[YH8000] “Configuration>>System>>Date&Time”
6.9.3
User Password Setting
Change the user password for when entering the setting screen from the YH8000. Enter the
current password and then enter a new password twice for confirmation.
Setup menu path:
[YH8000] “Configuration>>System>>Password”
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<6. Configuration>
6-13
6.9.4 Display
Configure settings related to the LU display and SCU display.
l Brightness adjustment of LU display
Set the brightness to any of 11 levels.
Setup menu path:
[HART] “Detailed setup>>System>>Local display>>LU LED display”
[YH8000] “Configuration>>System>>Local Display>>LU”
l Brightness adjustment of SCU display
Adjust backlight brightness and contrast to any of 11 levels.
Setup menu path:
[HART] “Detailed setup>>System>>Local display>>SCU LCD display”
[YH8000] “Configuration>>System>>Local Display>>SCU”
l Spectrum display of SCU display
Set whether or not to display the spectrum screen.
Setup menu path:
[HART] “Detailed setup>>System>>Local display>>SCU LCD display”
[YH8000] “CConfiguration>>System>>Local Display>>SCU”
Selection Option (HART
Display Name)
Hide
During alarm mode
Description
A spectrum is not displayed.
• Measurement spectrum
When any of the following alarms occur, the received optical signal and
absorption spectrum of the measured gas are displayed alternately in a
3-second cycle.
No.
Alarm
49
50
52
53
57
warning: detector signal high
fault: peak center out of range
warning: absorption too high
fault: transmission lost
fault: laser unit failure
• Reference cell spectrum
When any of the following alarms occur, the received optical signal and
absorption spectrum of the reference cell are displayed alternately in a
3-second cycle. (If the reference cell is disabled, the received optical signal
and absorption spectrum of the measured gas are displayed.)
No.
Alarm
51 fault: reference peak height low
54 fault: reference transmission low
55 fault: reference peak height high
Periodic
After display of the 6th display item on the screen is complete, the measurement
spectrum is displayed for 4 seconds.
Each measurement spectrum The received optical signal and absorption spectrum of the measured gas are
always displayed alternately in a 4-second cycle.
Each reference spectrum
The received optical signal and absorption spectrum of the reference cell are
always displayed alternately in a 4-second cycle.
The following shows the display image of each spectrum screen.
IM 11Y01D01-01EN
3rd Edition: Jun. 15, 2016-01
6-14
<6. Configuration>
Measurement absorption spectrum
Measurement received optical signal
Reference cell absorption spectrum
6.9.5
Reference cell received optical signal
Communication Address Setting
Configure the address settings for TDLS8000 and HART communication.
l IP address setting of TDLS8000
Setup menu path:
[HART] “Detailed setup>>System>>Communication>>TCP/IP>>Set IP settings”
[YH8000] “Configuration>>System>>Communication>>TCP/IP”
NOTE
When the IP address is changed, the TDLS8000 restarts automatically.
NOTE
When the IP address of the TDLS8000 is changed via the YH8000, the YH8000 connection
settings need to be configured again. For details, see “5.2.3 Connecting to the TDLS8000”.
l HART address setting
Setup menu path:
[HART] “Detailed setup>>System>>Communication>>HART output”
[YH8000] “Configuration>>System>>Communication>>HART”
“Loop current mode” is a setting related to multi-drop mode. For details on multi-drop mode, see
“7.5.1 Multidrop Mode”.
6.9.6
Moving Average Count for Analysis Values
Configure the setting for how many analysis cycles of spectrum data to take the moving average.
Increasing this value enables a more stable analysis result to be obtained but the response time
will increase. For explanations on the analysis cycle and average count, see “Appendix 5 What
Is an Analysis Period?”.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<6. Configuration>
6-15
Setup menu path:
[HART] “Detailed setup>>Analysis>>Averaging”
[YH8000] “Configuration>>Analysis>>Average”
6.9.7
Concentration Offset
It is possible to display the value resulting from adding a fixed offset value to the calculated
concentration value as the final reading. For the two-gas measurement specification, you can
set an offset for each of the first component gas concentration and second component gas
concentration.
Setup menu path:
[HART] “Detailed setup>>Analysis>>Zero offset”
[YH8000] “Configuration>>Analysis>>Zero Offset”
6.10
Non-process Parameter Settings
Non-process parameters are for correcting the influence on measurement values resulting from
the measurement target gas concentration contained in the purge gas when using gas containing
a component to be measured as the purge gas. Non-process parameters are used, for example,
with applications that use instrument air (approximately 20.9% oxygen) as the purge gas when
measuring the oxygen concentration of the process. This is because the measurement values
need to be corrected in consideration of the influence of laser absorption resulting from the
oxygen in the purge gas. The oxygen in the instrument air exists in various sections of the optical
path including the following.
• Laser module section in the laser unit (LU)
• Online validation cell in the sensor control unit (SCU)
• Alignment flanges (LU side and SCU side)
• Process insulation flanges (LU side and SCU side) (only when used)
• Insertion tubes (LU side and SCU side) (only when used)
• Valves (LU side and SCU side) (only when used)
• Process flanges (LU side and SCU side)
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
6-16
<6. Configuration>
General purpose and zone 2
Window of Isolation flange
Window of
Alignment flange
Insertion
Tube
Process
TDLS8000
LU or SCU
Optics
Module
A
B
C D
E
Non-process optical prth length
Non-Process Optical Prth Length
when with insertion tube
zone 1 model
Window of Isolation flange
Window for
Window of
Insertion
exp. construction
Alignment flange
tube
Process
TDLS8000
LU or SCU
Optics
Module
A1
A2
B
CD
E
Non-process optical prth length
Non-Process Optical Prth Length
when with insertion tube
Figure 6.2
6.10.1
Definitions of non-process optical path length
Non-process Optical Path Length
Set the optical path length of the non-process area.
Setup menu path:
[HART] “Detailed setup>>Analysis>>Non process param>>No-prcs OPL”
[YH8000] “Configuration>>Analysis>>Non-Process Parameters>>Path Length”
The length of the non-process optical path is the total of the lengths of the laser unit (LU) side and
sensor control unit (SCU) side. Calculate the length according to the corresponding procedure
below because it differs depending on the installation conditions.
l Non-process optical path length
The total target area differs depending on the installation conditions as shown in Table 6.1.
Furthermore, the length of each area of the total differs depending on the model name and code
as shown in Table 6.2 to Table 6.5. Use the values in Table 6.2 to Table 6.6 to calculate the nonprocess optical path length using the corresponding calculation formula in Table 6.1.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
6-17
<6. Configuration>
Table 6.1
Installation conditions
Installation conditions
LU section, SCU section and alignment flange only
When using insertion tube
When using process insulation flange
When using process insulation flange and insertion tube
Calculation formula
A + B + A’ + B’
A + B + E + A’ + B’ + E’
A + B + C + D + A’ + B’ + C’ + D’
A + B + C + D + E + A’ + B’ + C’ + D’ + E’
Note: In case of Zone 1 model, value A is A1 + A2, value A’ is A’1 + A’2.
Table 6.2
A, and A’ value by Gas parameter (General purpose and zone 2 model)
Model name and code
Optics Accessory, A (mm) A’ (mm)
Option Code
LU side SCU side
Model
Type
Gas parameter
other than -LA, /D
149.8
130.6
-X1
-LA
142.4
235.6
/D
142.4
130.6
other than -LA, /D
149.8
130.6
-X2
-LA
142.4
235.6
/D
142.4
130.6
-C1
72.2
92.6
other than -LA, /D
72.2
92.6
-C2
-LA
72.2
225.6
/D
72.2
92.6
other than -LA, /D
72.2
92.6
-G1
-C3
-LA
72.2
225.6
-G2
/D
72.2
92.6
-D2
TDLS8000
-C2
other than -LA, /D
72.2
92.6
-S2
-C4
-LA
72.2
225.6
-E2
/D
72.2
92.6
-C5
72.2
92.6
-A1
72.2
92.6
-A2
72.2
92.6
-D1
72.2
92.6
-D5
72.2
92.6
-H1
151.2
132.6
-H3
152.2
132.6
-H4
150.4
130.6
-S1
72.2
92.6
-L1
72.2
92.6
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<6. Configuration>
Table 6.3
6-18
A and A’ value by Gas parameter (zone 1 model)
Model name and code
Optics Accessory, A1+A2 (mm) A1'+A2' (mm)
Option Code
LU side
SCU side
Model
Type
Gas parameter
other than -LA, /D
138.6
119.4
-X1
-LA, /D
other than -LA, /D
138.6
119.4
-X2
-LA, /D
-C1
61.0
81.4
other than -LA, /D
61.0
81.4
-C2
-LA, /D
other than -LA, /D
61.0
81.4
-C3
-LA, /D
-D1
other than -LA, /D
61.0
81.4
-C1
-C4
-LA, /D
TDLS8000
-S1
-E1
-C5
61.0
81.4
-J1
-A1
61.0
81.4
-A2
61.0
81.4
-D1
61.0
81.4
-D5
61.0
81.4
-H1
140.0
121.4
-H3
141.0
121.4
-H4
139.2
119.4
-S1
61.0
81.4
-L1
61.0
81.4
Table 6.4
B and B’ value by Optics accessory
Model name and code
B (mm) B' (mm)
Optics Accessory
LU side SCU side
Model
Type
-NN
0
0
-G1
-LA
89.7
174.6
-G2
-U2
85.4
85.4
-D2
-C2
-U3
90.2
90.2
-S2
-U4
89.7
89.7
-E2
TDLS8000
-D5
83.9
83.9
-D1
-D8
85.9
85.9
-C1
-S1
-J5
81.9
81.9
-E1
-J8
83.9
83.9
-J1
-FC
83.4
83.4
Table 6.5
C, C’ and D, D’ value
Model name and code C, C’ (mm) D, D’ (mm)
IF8000
-21
-23
-31
-33
-41
-50
-80
-J5
-J8
-21
-21
-21
-21
-21
-50
-50
-50
-50
14
14
14
14
14
15.1
15.1
15.1
15.1
19.1
19.1
19.1
19.1
18.6
20.1
20.1
19.1
19.1
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
6-19
<6. Configuration>
Table 6.6
Insertion tube length
Part name
Insertion tube
E, E’ (mm)
Insertion tube length including flange part
(Depends on measurement application)
l Non-process optical path length of TDLS8000
Convert the LU section length + SCU section length [mm] to the unit set for the optical path length
and set the result as the non-process optical path length.
6.10.2
Non-process Pressure Setting
Set the pressure value of the non-process area.
Setup menu path:
[HART] “Detailed setup>>Analysis>>Non process param>>No-prcs pres”
[YH8000] “Configuration>>Analysis>>Non-Process Parameters>>Pressure”
6.10.3
Non-process Temperature Setting
Set the temperature input mode of the non-process area. Also, set the fixed temperature value
when the mode is fixed and offset value when the mode is “Active ambient.”
Setup menu path:
[HART] “Detailed setup>>Analysis>>Non process param>>No-prcs temp”
[YH8000] “Configuration>>Analysis>>Non-Process Parameters>>Temperature”
6.10.4
Non-process Concentration Setting
Set the concentration of the gas to be measured that is included in the non-process area.
Setup menu path:
[HART] “Detailed setup>>Analysis>>Non process param>>No-prcs conc”
[YH8000] “Configuration>>Analysis>>Non-Process Parameters>>Concentration”
6.11
Initializing the Settings (Factory Default
Settings)
This section describes the procedure to restore the settings to the state at the time of shipment
and lists the initial values of the parameters at the time of shipment.
6.11.1
Initialization Procedure
Execution menu path:
[HART] “Detailed setup>>System>>Initialize config”
[YH8000] >>Configuration>>System>>Configuration Initialization"
To execute initialization, open the above menu and then select from the following depending on
the types of parameters you wish to initialize. Multiple items can be selected at the same time.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
6-20
<6. Configuration>
Item name
(HART)
Setting data
AI/AO cal data
Calibration data
User info
Initialization target
All parameters settable from YH8000 and HART
(All parameters in section 6.11.2 except for “User info.”)
Input/output calibration data of AI/AO
Zero/span calibration data
TDLS8000 tag, IP settings, HMI (YH8000) user password, HART address,
HART 8-character short tag, write protect password
NOTE
When initialization is executed, the TDLS8000 restarts automatically.
6.11.2
Parameter Initial Value List
The initial values of parameters at the time of shipment are as follows.
Parameters with ○ in the “User specification” column are initialized to the values specified by the
customer at the time of ordering if the values were specified by the customer at that time.
l Process parameters
Parameter
OPL (Optical Path Length)
Pressure mode
Pressure value for fixed mode
Pressure active input source
Pressure value at 4mA
Pressure value at 20mA
Pressure backup mode when AI-1 input is out of range or under AI
calibration
Pressure backup set value
Temperature mode
Temperature value for fixed mode
Temperature active input source
Temperature value at 4mA input
Temperature value at 20mA input
Temperature backup mode when AI-2 input is out of range or
under AI calibration
Temperature backup set value
Temperature offset for active ambient method
Initial value
0.660[m]
Active input
101.325[kPa]
AI-1
40[kPa]
200[kPa]
Backup value
User
Min. – Max.
specification
○
0.01 – 100[m]
Select in the screen
0.1 – 10,000[kPa]
Select in the screen
0 – 10,000[kPa]
0 – 10,000[kPa]
Select in the screen
101.325[kPa]
Active input
25[deg C]
AI-2
0[deg C]
100[deg C]
Backup value
0.1 – 10,000[kPa]
Select in the screen
-273 – 3,000[deg C]
Select in the screen
-273 – 3,000[deg C]
-273 – 3,000[deg C]
Select in the screen
25[deg C]
-6[deg C]
-273 – 3,000[deg C]
-100 – 100[deg C]
l Non-process parameters
Parameter
Non-process OPL
Non-process pressure
Non-process temperature mode
Non-process temperature
Non-process temperature compensation coefficient for active
ambient method
Non-process concentration for measurement gas 1
Non-process concentration for measurement gas 2
Initial value
0[m]
101.325[kPa]
Fixed
25[deg C]
0.5
Min. – Max.
0 – 10[m]
0.1 – 10,000[kPa]
Select in the screen
-273 – 3,000[deg C]
0–1
0[ppm]
0[ppm]
0 – 1E6[ppm]
0 – 1E6[ppm]
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<6. Configuration>
6-21
l Units
Parameter
OPL unit
Pressure unit
Temperature unit
Initial value
m
kPa
deg C
User
Min. – Max.
specification
○
Select in the screen
○
Select in the screen
○
Select in the screen
l Process alarms
Parameter
Warning selection
Transmission low warning limit
Pressure low warning limit
Pressure high warning limit
Temperature low warning limit
Temperature high warning limit
Gas1 concentration low warning limit
Gas1 concentration high warning limit
Gas2 concentration low warning limit
Gas2 concentration high warning limit
Initial value
All selected
20[%]
90[kPa]
110[kPa]
0[deg C]
100[deg C]
10[ppm]
900,000[ppm]
10[ppm]
900,000[ppm]
Min. – Max.
Select in the screen
0 – 100[%]
0.1 – 10,000[kPa]
0.1 – 10,000[kPa]
-273 – 3,000[deg C]
-273 – 3,000[deg C]
0 – 1E6[ppm]
0 – 1E6[ppm]
0 – 1E6[ppm]
0 – 1E6[ppm]
l Analog output
Parameter
AO output item
Measurement value at 4mA output
Measurement value at 20mA output
AO hold mode during warning
Preset hold value during warning
Preset hold delay during warning
AO hold mode during fault
Preset hold value during fault
Preset hold delay during fault
AO hold mode during calibration/validation
Preset hold value during calibration/validation
AO hold mode during maintenance
Preset hold value during maintenance
AO hold mode during warm-up
Preset hold value during warm-up
Initial value
Concentration 1
Concentration 0[ppm]
Transmission 0[%]
Temperature 0[deg C]
Pressure
0[kPa]
Concentration 100[ppm]
Transmission 100[%]
Temperature 100[deg C]
Pressure
0.1[kPa]
Non-Hold
3.0[mA]
0
Preset hold
3.0[mA]
0
Preset hold
3.8[mA]
Preset hold
3.8[mA]
Preset hold
3.8[mA]
User
Min. – Max.
specification
○
Select in the screen
○
-1E7 – 1E7[ppm]
-1E7 – 1E7[%]
-1E7 – 1E7[deg C]
-1E4 – 1E4[kPa]
○
-1E7 – 1E7[ppm]
-1E7 – 1E7[%]
-1E7 – 1E7[deg C]
-1E4 – 1E4[kPa]
Select in the screen
3.0 – 21.6[mA]
0–5
Select in the screen
3.0 – 21.6[mA]
0–5
Select in the screen
3.8 – 20.5[mA]
Select in the screen
3.8 – 20.5[mA]
Select in the screen
3.8 – 20.5[mA]
l Digital output
Parameter
Number of output delays for warning and fault
DO output item selection
Initial value
0
All selected
Min. – Max.
0 – 100
Select in the screen
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<6. Configuration>
6-22
l Digital input
Parameter
Initial value
0.5[s]
Disable
Min. – Max.
Select in the screen
Select in the screen
Initial value
Cal/Val
Stream 1
Disable
60[min]
Min. – Max.
Select in the screen
Select in the screen
Select in the screen
1 – 1,440[min]
Parameter
Initial value
30[min]
Min. – Max.
Select in the screen
Parameter
Initial value
Filter time
DI item selection
l Valve
Parameter
Valve usage
Initial stream
Next stream of stream switching by time duration
Time duration of next stream
l Loop check
Test auto release time
l System
User averaging number
Analyzer tag
Analyzer IP address
Subnet mask
Default gateway address
HMI user password
LCD spectrum display mode
LCD backlight brightness
LCD contrast
LED brightness
*1:
1
Blank
192.168.1.10
255.255.255.0
192.168.1.254
1234
Hide
10
5
10
User
specification
○
Min. – Max.
1 – 32(*1)
ASCII 32 characters
IPv4 address
IPv4 address
IPv4 address
ASCII 8 characters
Select in the screen
0 – 10
0 – 10
0 – 10
The maximum value varies depending on the measurement target gas (application). Normally, it is 16.
l HART parameters
Parameter
Polling address
Loop current mode
Write protect password
Short tag
TV item
QV item
Response preamble number
Memo 1/ Memo 2/ Memo 3
Message
Descriptor
Trim Who/Trim Desc/Trim Loc
Date/Trim date
Initial value
0
Enable
All space
All space
Temperature
Transmission
5
All space
All space
All space
All space
1900/01/01
Configuration locked status mask
Off
*1:
Min. – Max.
0 – 63
Select in the screen
ASCII 8 characters
8 characters (*1)
Select in the screen
Select in the screen
5 – 20
ASCII 32 characters
32 characters (*1)
16 characters (*1)
ASCII 16 characters
1900/01/01 –
2155/12/31
Select in the screen
Uppercase letters, numbers, and symbols can be entered.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<6. Configuration>
6-23
l Zero calibration
Parameter
Auto zero calibration time initiate
Auto zero calibration time initiate cycle (day)
Auto zero calibration time initiate cycle (hour)
Auto zero calibration time initiate base clock
Initial value
Disable
0 (=Disable)
0 (=Disable)
2010/01/01 00:00:00
Auto zero calibration gas purge time
Auto zero calibration process purge time
Auto valve control for manual zero calibration
600[s]
600[s]
Disable
Min. – Max.
Select in the screen
0 – 999
0 – 23
2010/01/01 00:00:00 –
2068/12/31 23:59:59
0 – 10,000[s]
0 – 10,000[s]
Select in the screen
l Span calibration
Parameter
Pressure mode for span calibration
Temperature mode for span calibration
OPL mode for span calibration
Pressure fixed value for span calibration
Temperature fixed value for span calibration
OPL fixed value for span calibration
Gas1 concentration value for span calibration
Gas2 concentration value for span calibration
Gas type for span calibration
Auto span calibration time initiate
Auto span calibration time initiate cycle (day)
Auto span calibration time initiate cycle (hour)
Auto span calibration time initiate base clock
Initial value
Process parameter
Process parameter
Process parameter
101.325[kPa]
25[deg C]
0.66[m]
219,000[ppm]
219,000[ppm]
Gas 1
Disable
0 (=Disable)
0 (=Disable)
2010/01/01 00:00:00
Auto span calibration gas purge time
Auto span calibration process purge time
Auto valve control for manual span calibration
600[s]
600[s]
Disable
*1:
Min. – Max.
Select in the screen
Select in the screen
Select in the screen
0.1 – 10,000[kPa]
-273 – 3,000[deg C]
0.01 – 100[m]
0 – 1E6[ppm](*1)
0 – 1E6[ppm](*1)
Select in the screen
Select in the screen
0 – 999
0 – 23
2010/01/01 00:00:00 –
2068/12/31 23:59:59
0 – 10,000[s]
0 – 10,000[s]
Select in the screen
Zero is not allowed.
l Zero + span calibration
Parameter
Auto zero + span calibration time initiate
Auto zero + span calibration time initiate cycle (day)
Auto zero + span calibration time initiate cycle (hour)
Auto zero + span calibration time initiate base clock
Initial value
Disable
0 (=Disable)
0 (=Disable)
2010/01/01 00:00:00
Min. – Max.
Select in the screen
0 – 999
0 – 23
2010/01/01 00:00:00 –
2068/12/31 23:59:59
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
6-24
<6. Configuration>
l Offline validation (*1)
Parameter
Pressure mode for offline validation 1
Temperature mode for offline validation 1
OPL mode for offline validation 1
Pressure fixed value for offline validation 1
Temperature fixed value for offline validation 1
OPL fixed value for offline validation 1
Gas1 concentration value for offline validation 1
Gas2 concentration value for offline validation 1
Gas type for offline validation 1
Auto offline validation 1 time initiate
Auto offline validation 1 time initiate cycle (day)
Auto offline validation 1 time initiate cycle (hour)
Auto offline validation 1 time initiate base clock
Auto offline validation 1 gas purge time
Auto offline validation 1 process purge time
Auto valve control for manual offline validation 1
Auto offline validation 1+2 time initiate
Auto offline validation 1+2 time initiate cycle (day)
Auto offline validation 1+2 time initiate cycle (hour)
Auto offline validation 1+2 time initiate base clock
*1:
*2:
Initial value
Process parameter
Process parameter
Process parameter
101.325[kPa]
25[deg C]
0.66[m]
200,000[ppm]
200,000[ppm]
Gas 1
Disable
0 (=Disable)
0 (=Disable)
2010/01/01 00:00:00
Min. – Max.
Select in the screen
Select in the screen
Select in the screen
0.1 – 10,000[kPa]
-273 – 3,000[deg C]
0.01 – 100[m]
0 – 1E6[ppm](*2)
0 – 1E6[ppm](*2)
Select in the screen
Select in the screen
0 – 999
0 – 23
2010/01/01 00:00:00 –
2068/12/31 23:59:59
600[s]
0 – 10,000[s]
600[s]
0 – 10,000[s]
Disable
Select in the screen
Disable
Select in the screen
0 (=Disable)
0 – 999
0 (=Disable)
0 – 23
2010/01/01 00:00:00 2010/01/01 00:00:00 –
2068/12/31 23:59:59
The initial value for the parameters related to “offline validation 2” is the same as those for “offline validation 1.”
Zero is not allowed.
l Online validation (*1)
Parameter
Temperature mode for online validation 1
Temperature ambient offset for online validation 1
Pressure value for online validation 1
Temperature fixed value for online validation 1
OPL value for online validation 1
Gas1 concentration value for online validation 1
Gas2 concentration value for online validation 1
Gas type for online validation 1
Auto online validation 1 time initiate
Auto online validation 1 time initiate cycle (day)
Auto online validation 1 time initiate cycle (hour)
Auto online validation 1 time initiate base clock
Initial value
Active ambient
-2.2[deg C]
101.325[kPa]
25[deg C]
0.1306[m](*2)
200,000[ppm]
200,000[ppm]
Gas 1
Disable
0 (=Disable)
0 (=Disable)
2010/01/01 00:00:00
Auto online validation 1 validation gas purge time
Auto online validation 1 normal gas purge time
Auto valve control for manual online validation 1
Reading mode for online validation 1
Concentration output factor during online validation 1
600[s]
600[s]
Disable
Process+Validation
1.0
*1:
*2:
Min. – Max.
Select in the screen
-100 – 100[deg C]
0.1 – 10,000[kPa]
-273 – 3,000[deg C]
0.01 – 10[m]
-1E6 – 1E6[ppm]
-1E6 – 1E6[ppm]
Select in the screen
Select in the screen
0 – 999
0 – 23
2010/01/01 00:00:00 –
2068/12/31 23:59:59
0 – 10,000[s]
0 – 10,000[s]
Select in the screen
Select in the screen
-9.9 – 9.9
The initial value for the parameters related to “online validation 2” is the same as those for “online validation.”
The parameter may be initialized to a value different from the one shown because the initial value differs depending on the
measurement target gas (application) of the TDLS8000.
l Concentration offset
Parameter
Concentration offset for gas1
Concentration offset for gas2
Initial value
0[ppm]
0[ppm]
Min. – Max.
-1E6 – 1E6[ppm]
-1E6 – 1E6[ppm]
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
7.
7-1
<7. HART Communication>
HART Communication
The following functions can be performed via HART communication.
• Checking concentration, transmission, process pressure, and process temperature
• Checking alarm statuses
• Setting parameters
• Performing calibration and validation
• Performing loop check
• Checking alarms and calibration history
This chapter explains matters specific to HART communication.
7.1 Connection
For the method of connecting a HART setting tool to the TDLS8000, see “5.1
HART Configuration Tool”.
7.2
Connecting the
Menu Tree
This section shows the hierarchal configuration of the DD menu. For the whole configuration
containing all parameters, see “Appendix 3
General View of HART DD”.
7.2.1
DD Menu
The root menu (top menu) is as follows.
Root menu
Process variables
Diagnosis/Service
Basic setup
Detailed setup
Review
Description
Displays the latest values of PV-QV and measurements.
Checks alarms and history; performs calibration, validation, and loop check
Assigns PV-QV items; sets output ranges
Sets parameters specific to the TDLS8000
Displays measurements, I/O values, and manufacturing information
The menus in the second and lower layers are as follows.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<7. HART Communication>
7-2
Menu
ルートメニュー
•
•
•
•
•
Process variables
Diagnosis/Service
Basic setup
Detailed setup
Review
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<7. HART Communication>
7-3
Menu
ルートメニュー
•
•
•
•
•
Process variables
Diagnosis/Service
Basic setup
Detailed setup
Review
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<7. HART Communication>
7-4
Menu
ルートメニュー
•
•
•
•
•
Process variables
Diagnosis/Service
Basic setup
Detailed setup
Review
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
7-5
<7. HART Communication>
7.2.2
DTM Menu (FieldMate)
FieldMate is Yokogawa’s HART setting tool that runs on PCs. The menu displayed on FieldMate
slightly differs from the DD menu.
As shown in the figure below, the root menu (top menu) of FieldMate is “DTM Menu (Online)”,
under which the DD menu comes as five sub-menus (see 7.2.1). The configuration under the
sub-menus is the same as that of the DD menu.
DTM Menu(Online)
• Device
Configuration
• Device setup
• Diagnostic
• Diagnostic/Service
• Process Variable
• Process variable
• Maintenance
• Maintenance
*1:
7.3
• Basic setup
• Detailed setup
• Review
(*1)
Of the commands in the “Diagnostic/Service” menu, the execution commands for calibration, validation, and loop check are
included.
Write Protection
The write protection via HART communication is a dual system: a hardware switch and software
certification. When both protections are disabled, data can be written to the TDLS8000.
The current protection status can be confirmed from the value of “Yes” or “No” of the “Write
protect” parameter, which is retrieved by either of the following menus.
• “Review>>Field device info”
• “Detailed setup>>Field device info>>Write protect menu”
7.3.1
Hardware Write Protection
By turning the switch ON on the SCU, write protection is enabled. The default setting is OFF.
Switch position Write protection
OFF
Disabled
ON
Enabled
HART WP
OFF
ON
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
7.3.2
7-6
<7. HART Communication>
Software Write Protection
By setting a password (eight alphanumeric characters), write protection is enabled. The
password can be set and changed on the “New password” screen. When a password has been
set and it is entered in the “Enable write 10min” screen, the protection is disabled for 10 minutes.
When any data is written to the TDLS8000, the disabled time will be extended for another 10
minutes. This means that the protection is enabled 10 minutes after the last writing.
To disable software write protection indefinitely, enter 8 spaces in the “New password” screen. As
the default, protection is disabled (8 spaces have been entered).
l How to set a password
How to set a password will be explained using the character string “12345678” as an example.
(1) Go down the menu to Write protect menu: “Detailed setup>>Field device info>>Write
protect menu”. The screen shown below will appear. Select the “New password” line.
(2) Enter “12345678” and click “ENTER”.
(3) Confirm that “12345678” has been entered and click ENTER within 30 seconds.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
7-7
<7. HART Communication>
(4) When the new password is set, the following screen will appear. Click “OK”.
NOTE
When the TDLS8000 is in either of the following statuses, writing via HART communication is not
allowed even if write protection is disabled.
• A password for maintenance has been entered on the YH8000 and the setting change is
enabled (during maintenance)
• Calibration and validation are being performed.
l How to disable protection for 10 minutes
(1) Go down the menu to Write protect menu: “Detailed setup>>Field device info>>Write
protect menu”. The following screen will appear. Select the “Enable write 10min” line.
(2) Enter “12345678” and click “ENTER”.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<7. HART Communication>
7-8
(3) After successful verification, the following screen will appear. Click “OK”. Click “OK” again
on the next screen.
7.3.3
Device Configuration Locked
The TDLS8000 in the write protection status is recognized by some host systems as being in the
“Device Configuration Locked” status, which is defined in HART communication, indicating that
the TDLS8000 is in an abnormal condition. A window may pop up on a field communicator to
prompt confirmation.
To avoid this, the TDLS8000 has a function to mask this status. By setting this mask to ON,
the “Device Configuration Locked” status will not occur even in the write protection status. The
default setting is OFF. The setting will be maintained even after the power is turned off.
Menu: “Diagnosis/Service>>Test/Status>>Status mask>>Dev cfg locked mask”
7.4
Alarm Definition (Status group)
This section explains the device-specific alarms on HART communication and their definition.
On a HART setting tool, device-specific alarms and their status information are bundled in groups
consisting of up to eight items. These groups are defined as “Status group#” and alarms are
expressed in the format of the character string of an alarm followed by (AL-alarm number). The
details of each group are given below.
Display menu: “Diagnosis/Service>>Test/Status>>View status”
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
7-9
<7. HART Communication>
Group
Group 1
Group 2
Group 3
Group 4
Group 6
Group 7
Group 8
Group 9
Group 10
Status
Transmission Low (AL-01)
Pressure Low (AL-02)
Pressure High (AL-03)
Temperature Low (AL-04)
Temperature High (AL-05)
Conc Gas1 Low (AL-06)
Conc Gas1 High (AL-07)
Conc Gas2 Low (AL-08)
Conc Gas2 High (AL-09)
LU Temp Low (AL-10)
LU Temp High (AL-11)
SCU Temp Low (AL-12)
SCU Temp High (AL-13)
Validation Required (AL-14)
Validation Error (AL-15)
Zero Cal Error (AL-16)
Span Cal Error (AL-17)
Non Process Alarm (AL-19)
AI-1 (Pres) Low (AL-20)
AI-1 (Pres) High (AL-21)
AI-2 (Temp) Low (AL-22)
AI-2 (Temp) High (AL-23)
External Alarm (AL-24)
Clock Adj Required (AL-25)
Setting File Error (AL-26)
Calib File Error (AL-27)
Laser Md Temp Low (AL-45)
Laser Md Temp High (AL-46)
Laser Temp Low (AL-47)
Laser Temp High (AL-48)
Detect Signal High (AL-49)
Peak Center OOR (AL-50)
Ref Peak Height Low (AL-51)
Absorption High (AL-52)
Transmission Lost (AL-53)
Ref Trans Low (AL-54)
R Peak Height High (AL-55)
Outlier Reject Lmt (AL-56)
Laser Unit Fail (AL-57)
Inter Comm Fail (AL-58)
Laser Module Error (AL-59)
File Access Error (AL-60)
EEPROM Error (AL-61)
LU Connect Error (AL-62)
FPGA Failure (AL-63)
Program Error (AL-64)
Warm-up
Maintenance mode
Zero Cal
Span Cal
Offline Val
Online Val
AI-1 (Pres) Cal
AI-2 (Temp) Cal
Attribute
Warning
Description
See “10.2 Warning Display and
Handling”.
Fault
See “10.1 Fault Display and
Handling”.
Warning
Fault
See “10.2 Warning Display and Handling”.
See “10.1 Fault Display and
Handling”.
Warning
Fault
See “10.2 Warning Display and Handling”.
See “10.1 Fault Display and
Handling”.
Warning
Fault
See “10.2 Warning Display and Handling”.
See “10.1 Fault Display and
Handling”.
Status
Warming-up
Maintenance
Zero calibration
Span calibration
Off-line validation
Online validation
AI-1 (pressure input) calibration
AI-2 (temperature input) calibration
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
7-10
<7. HART Communication>
7.5
Functions Specific to HART
Communication
This section explains functions that can be performed only via HART communication. These
functions include those specified by HART communication and those of the TDLS8000 only for
HART communication.
7.5.1
Multidrop Mode
In the multidrop mode, multiple HART communication devices can be connected to a single
HART communication line. Set “Poll addr” to a value of 0 to 63 so that each device has a different
address.
Set “Loop current mode” to “Disabled”. In this setting, AO-1 output will be fixed to 4 mA, and
burnout output will be disabled. However, in the multidrop connection with devices that receive
analog outputs (including actuators), one unit in one loop can output analog signals. In this case,
set “Loop current mode” to “Enabled”.
Configuration menu: “Detailed setup>>System>>Communication>>HART output”
7.5.2 Squawk
This function identifies any unit of the TDLS8000 that is connected via HART communication.
“SQUAWK !!!” will appear on the SCU display of the TDLS8000 as shown below.
l How to use the Squawk function
(1) Go down the menu to Squawk: “Diagnosis/Service>>Loop check>>Squawk”.
(2) Choose an operation of Squawk among the following.
Operation
Disables Squawk.
Enables Squawk.
Continues the display until “Off” is chosen or the TDLS8000 is turned off.
Squawk Once Enables Squawk for 10 seconds.
Off
On
7.5.3
Option
Aborting Calibration and Validation
Even if a setting tool is accidentally disconnected during calibration or validation via HART
communication, these tasks do not suspend but continue on the TDLS8000. In this case, stop
calibration and validation with the abort function, and then start these tasks afresh. Note that
calibration and validation commanded from the YH8000 cannot be aborted with this function.
The abort function is available for calibration and validation.
For calibration: “Diagnosis/Service>>Calibration>>Abort calibration”
For validation: ”Diagnosis/Service>>Validation>>Abort validation”
The following shows how to abort manual span calibration.
(1) During the span calibration, go down the menu to Abort calibration:
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
7-11
<7. HART Communication>
“Diagnosis/Service>>Calibration>>Abort calibration”
(2) After clicking “OK” on the warning screen, the screen shown below will appear.
If calibration is not being performed, an error message will appear. Confirm that the abortion
target is span calibration, and select “Yes”.
(3) The process gas injection screen will appear.
When automatic valve control is disabled, manually open the valve and purge the calibration
cell with process gas. After confirming that the process gas concentration has stabilized,
click “OK”.
(4) The screen shown below appears and the abortion of span calibration will finish.
Click “OK” to return to the menu.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
Blank Page
8.
8-1
<8. YH8000 HMI Unit>
YH8000 HMI Unit
The YH8000 HMI unit is an HMI (Human Machine Interface) option for the TDLS8000. This
chapter explains how to use the YH8000.
CAUTION
To prevent damaging the touch panel, do not use pointed objects (e.g., ballpoint pen, pencil),
objects with narrow tips, hard objects, and the like when you use the touch panel.
Also, to prevent erroneous operation, touch the center of controls.
Avoid applying strong load on the screen to prevent damage or malfunction.
8.1 Connection
For details on how to connect the YH8000 HMI unit to the TDLS8000 Tunable Laser
Spectrometer, see “5.2 Connecting to the YH8000”.
8.2
Home Screen
The home screen shown in Figure 8.1 or Figure 8.2 is the YH8000 main screen. Tapping
the bottom of the screen shows the home screen.
at
If a TDLS8000 is not connected to the YH8000, switching to the home screen is not possible.
If multiple TDLS8000s are connected, you can select to show the overall display. For details on
the overall display, see “8.7.1 Overall Display”.
Figure 8.1
Home screen (when a single TDLS8000 is connected)
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
8-2
<8. YH8000 HMI Unit>
Figure 8.2
8.2.1
Home screen (overall display when three TDLS8000s are connected)
Home Screen Display Items
Date and time Operating status
Tag name
HART communication status
Concentration
Concentration meter
Transmission meter
Transmission
Process pressure
Process temperature
Pressure icon
Temperature icon
Buttons at the bottom of the screen
Figure 8.3
Home screen display items
Date and time
Shows the date and time set on the TDLS8000.
Tag
Shows the TDLS8000 tag name.
If the overall display is shown, “Overall” is displayed.
Operating status
Shows the TDLS8000 operating status.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
8-3
<8. YH8000 HMI Unit>
Display
Measuring
Warm-up
Maintenance
Zero Calibration
Span Calibration
Offline Validation
Online Validation
Meaning
Measuring (normal operation)
Warming up
Maintenance in progress
Zero calibration in progress
Span calibration in progress
Offline validation in progress
Online validation in progress
HART communication in progress
An icon appears when a HART command is received.
Concentration
Shows the concentration.
If any of the following alarms are occurring, “***” is displayed because the concentration
cannot be calculated.
Alarm
number
49
50
52
53
56
Alarm name
Detector Signal High
Peak Center Out of Range
Absorption too High
Transmission Lost
Outlier Rejection Limit
Concentration meter
Shows the concentration with a meter.
Abnormal range (white)
Normal region (gray)
Abnormal range (white)
Right end value (max.)
Left end value (min.)
The white area indicates outside the concentration high/low limit alarm range. For details on
high/low concentration limit alarm, see “6.6 Process Alarm Settings”. For details on how to
set the meter display range, see “8.2.3
Setting the Meter Range”.
Transmission
Shows the laser beam transmission.
Transmission meter
Shows the transmission with a meter.
Abnormal range (white)
Left end value (min.)
Normal region (gray)
Right end value (max.)
The white area indicates outside the transmission low limit alarm range. For details on
transmission low limit alarm, see “6.6
Process Alarm Settings”. The display range is
fixed to 0 to 100% and cannot be changed.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
8-4
<8. YH8000 HMI Unit>
Process temperature and process pressure
Shows the process temperature and process pressure.
Temperature icon and pressure icon
Shows the process temperature and process pressure input modes.
: Fixed value
: Non-fixed value
Buttons at the bottom of the screen
Selects different screens.
Button
Button name
Home button
Displays the home screen.
Trend display
button
Alarm
information
button
Displays the trend screen. You can view the measurement
trends of multiple TDLS8000s on the same screen.
Configuration
display button
Configuration
button
8.2.2
Description
Displays the TDLS8000 alarm screen.
Displays the current TDLS8000 settings. The button can
also be used to display I/O values, alarm history, and so
on.
Executes TDLS8000 configuration, calibration, and the
like. It can also be used to change the YH8000 connection
destination, IP settings, and so on.
Selecting the Style
There are two home screen display styles: Bar meter and Arc meter. Select the style of your
liking. The factory default setting is Bar meter.
Figure 8.4
Bar meter style
Arc meter style
Configuration menu path:
“
8.2.3
>>HMI>>Display Setting>>Home Style”
Setting the Meter Range
You can set the concentration meter display range. The available range and default values are
shown in the following table.
Left end value (min.)
Right end value (max.)
Lower limit
0%
0ppm
Left end value
Upper limit
Right end value
100%
1000000ppm
Default value
0%
0ppm
100%
10000ppm
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
8-5
<8. YH8000 HMI Unit>
Configuration menu path:
“
>>HMI>>Display Setting>>Meter Range”
NOTE
Changing the meter display range will not change the analog output range of the TDLS8000.
8.2.4
Alarm Indicator
If an alarm is occurring on the TDLS8000, an icon is displayed to the left of the tag name at the
top of the screen and to the right of
at the bottom of the screen. Also, the area related to the
alarm is enclosed in a frame.
These alarm indicators remain blinking until you check the information on the alarm screen.
The indicators also start blinking when a new alarm occurs. If the alarm ceases, the alarm icons
disappear.
There are two types of alarm icons: a red icon indicating a fault (occurs when the various types of
diagnostic information being monitored by the self-diagnostics function are clearly abnormal and
correct concentration calculation is not possible) and a yellow icon indicating a warning (occurs
when the various types of diagnostic information being monitored by the self-diagnostics function
are outside the normal range).
Alarm icon types
Icon
Color
Meaning
Red
Fault
Yellow
Warning
Alarm icon to the left of the tag name
Concentration alarm indicator
Transmission alarm indicator
Temperature alarm indicator
Pressure alarm indicator
Alarm icon to the right of the alarm button
Figure 8.5
Alarm Indicator
Alarm icon to the left of the tag name
Indicates that an alarm is occurring on the applicable TDLS8000.
Alarm icon to the right of the alarm button
Indicates that an alarm is occurring on the connected TDLS8000.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
8-6
<8. YH8000 HMI Unit>
Concentration alarm indicator
Indicates that any of the following alarms is occurring.
Alarm
number
6
7
8
9
Alarm name
Concentration Gas1 Low
Concentration Gas1 High
Concentration Gas2 Low
Concentration Gas2 High
Transmission alarm indicator
Indicates that any of the following alarms is occurring.
Alarm
number
1
53
Alarm name
Transmission Low
Transmission Lost
Temperature alarm indicator
Indicates that any of the following alarms is occurring.
Alarm
number
4
5
Alarm name
Process Temperature Low
Process Temperature High
Pressure alarm indicator
Indicates that any of the following alarms is occurring.
Alarm
number
2
3
8.3
Alarm name
Process Pressure Low
Process Pressure High
Trend Screen
Tapping
at the bottom of the screen shows the trend screen. The trend screen displays
the trend waveforms of four items. The items that can be displayed include concentration,
transmission, process temperature, and process pressure.
If there is no connected TDLS8000, switching to the trend screen is not possible.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
8.3.1
8-7
<8. YH8000 HMI Unit>
Trend Screen Display Items
Item name
Current value
Configuration button
Figure 8.6
Update pause button
Trend Screen
Item name
Shows the item of the displayed trend waveform.
If multiple TDLS8000s are connected, analyzer numbers are added in front of item names
as shown below.
Current value
Shows the current value of the displayed item.
Configuration button
Tapping
at the bottom of the screen shows the trend configuration screen.
Update pause button
Tapping
updating.
pauses the updating of the trend waveforms. Tapping it again resumes the
Being updated <=>
Paused
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
8-8
<8. YH8000 HMI Unit>
8.3.2
Selecting the Items to Display
Press
on the trend screen to display the trend configuration screen.
Analyzer
number selection
Trend number
selection
Display item
selection
Figure 8.7
Trend configuration screen
(1) Select the trend number you want to change.
(2) Select a analyzer number. (When multiple TDLS8000s are connected)
(3) Select the display item.
You can select from the following display items for the trend waveforms.
Trend display items
Available option
Description
Gas1 Concentration Component 1 gas concentration
Gas1 STDEV
Standard deviation of component 1 gas concentration
Component 2 gas concentration (selectable only for two-gas
Gas2 Concentration
measurement)
Standard deviation of component 2 gas concentration
Gas2 STDEV
(selectable only for two-gas measurement)
Transmission
Transmission
Temperature
Process temperature
Pressure
Process pressure
None
Nothing
The factory default settings are shown in the following table.
Factory default display items
Trend number
Trend 1
Trend 2
Trend 3
Trend 4
Description
Gas1 Concentration
Transmission
Temperature
Pressure
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
8-9
<8. YH8000 HMI Unit>
8.3.3
Setting the Displayed Time
Press
on the trend screen to display the trend configuration screen.
Display time
span selection
Time axis
configuration
screen selection
Figure 8.8
Trend time axis configuration screen
(1) Switch to the time axis configuration screen.
(2) Set the display time span.
You can select the display time span of the trend graph from the available options in the following
table.
Trend display items
Available option Description
1 min
3 min
5 min
10 min
30 min
1 hour
3 hour
6 hour
12 hour
1 minute
3 minutes
5 minutes
10 minutes
30 minutes
1 hour
3 hours
6 hours
12 hours
The factory default setting is 30 min.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
8-10
<8. YH8000 HMI Unit>
8.3.4
Setting the Vertical Scale
Press
on the trend screen to display the trend configuration screen.
Trend number
selection
Vertical scale
setting
Vertical scale
Auto adjustment
execution
Figure 8.9
Trend configuration screen
(1) Select the trend number you want to change.
(2) Tapping Auto Scale automatically sets the most suitable value for that instant.
(3) To specify the scale manually, set the upper scale and lower scale.
8.4
Alarm Screen
Tapping
at the bottom of the screen shows the alarm screen. The alarm screen displays a list
of alarms that are currently occurring. For the meanings of alarms and correct actions, see “10.
Troubleshooting”.
NOTE
When multiple TDLS8000s are connected to the YH8000, tapping
displays a TDLS8000
selection screen first. Selecting the target TDLS8000 and touching OK will display the alarm
screen of the selected TDLS8000.
For details, see “8.7.2 TDLS8000 Selection Screen”.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<8. YH8000 HMI Unit>
Alarm icon
8-11
List line number Alarm number Alarm description Sub number
Scroll up a page
Scroll up a line
Scroll down a line
Scroll down a page
Figure 8.10
Alarm Screen
NOTE
The sub number is a number that indicates the details of the alarm. It is a number that Yokogawa
service representatives use for troubleshooting.
8.5
Information Screen
You can view various types of TDLS8000 information on the information screen.
Tapping
at the bottom of the screen shows the information screen.
NOTE
When multiple TDLS8000s are connected to the YH8000, tapping
displays a TDLS8000
selection screen first. Selecting the target TDLS8000 and touching OK will display the information
screen of the selected TDLS8000.
For details, see “8.7.2 TDLS8000 Selection Screen”.
Figure 8.11
Information Screen
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
8.5.1
8-12
<8. YH8000 HMI Unit>
I/O List Screen
The I/O list screen displays the status of analog I/O, digital I/O, valve control power output, and
SCU and LU temperatures. Tapping the I/O button on the
list screen.
Figure 8.12
8.5.2
information screen displays the I/O
I/O list
Configuration View Screen
On the configuration view screen, you can view the TDLS8000 settings. Tapping the
Configuration button on the
information screen displays the following menu. For details on
the menu tree, see under Configuration in “Appendix 4 YH8000 Menu Tree”.
Figure 8.13
Configuration View Screen
NOTE
You cannot change the settings on the configuration view screen. To change them, see “8.6
Configuration Screen”.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<8. YH8000 HMI Unit>
8.5.3
8-13
System Information Screen
The system information screen displays the TDLS8000’s tag name, serial number, IP address,
MAC address, HART device ID, software version, and analysis period and the laser module’s
serial number. Tapping the System Information button on the
system information screen.
Figure 8.14
8.5.4
information screen displays the
System Information Screen
Spectrum Screen
The spectrum screen displays the spectrum being measured. Tapping the Spectrum button on
the
information screen displays the spectrum screen.
Detector signal
Measurement absorption spectrum
Reference cell
detector signal
Reference cell
absorption spectrum
Spectrum capture
Figure 8.15
Spectrum Screen
Detector signal
Shows the detector signal on the measured gas side.
Measurement absorption spectrum
Shows the absorption spectrum of the measured gas.
Reference cell detector signal
Shows the detector signal of the reference cell. This can be shown only on a TDLS8000 with
a valid reference cell.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
8-14
<8. YH8000 HMI Unit>
Reference cell absorption spectrum
Shows the absorption spectrum of the reference cell. This can be shown only on a
TDLS8000 with a valid reference cell.
Spectrum capture
Records spectrum waveform data to the TDLS8000 internal storage. Do not use this under
normal circumstances. Use it only if you receive a request from your Yokogawa service
representative.
NOTE
The spectrum screen is used by Yokogawa service representatives for troubleshooting purposes.
You may be asked to check this screen depending on the situation. In such a case, please
operate the screen according to their instructions.
8.5.5
Alarm History Screen
The alarm history screen can be used to check alarms and messages that occurred in the past.
Tapping Log Book on the
history screen.
information screen and then Alarm History displays the alarm
The maximum number of history events that you can view on the YH8000 is 99. For details on
alarms, see “10.
Troubleshooting”. For details on messages, see “9.12 Alarm History”.
Alarm icon
Timestamp
Sub number
List line number
Alarm number
Status
Alarm description
Scroll up a page
Scroll up a line
Scroll down a line
Scroll down a page
Figure 8.16
Alarm History Screen
Alarm icon
Nothing is displayed for messages. While the alarm is active, the icon appears bright. When
the alarm is cleared, the icon changes to a dark indication.
Status
“On” is displayed when the alarm is active, and “Off” when it is cleared.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
8.5.6
8-15
<8. YH8000 HMI Unit>
Cal/Val History Screen
The Cal/Val history screen can be used to view the results of calibrations and validations that
have been executed in the past. Tapping Log Book on the
Val History displays the Cal/Val history screen.
information screen and then Cal/
The maximum number of history events that you can view on the YH8000 is 99.
List line number Timestamp
Cal/Val type
Result
Scroll up a page
Scroll up a line
Scroll down a line
Scroll down a page
Figure 8.17
Cal/Val History Screen
The result column displays the following information depending on the Cal/Val type.
Information displayed in Cal/Val history
Cal/Val type
Man Zero Cal
Auto Zero Cal
Man Span Cal
Auto Span Cal
Man Offline Val
Auto Offline Val
Man Online Val
Auto Online Val
Transmission
Zero Cal Restored
Span Cal Restored
Result
Successful: OK
Unsuccessful:
Error
Successful: OK
Unsuccessful:
Error
Pass
Fail
(Max Absorption=)
(Conc=Coef=)
(Exp=Rslt=)
Trans=
Restored to factory default settings: By Factory data
Restored to previous settings: By Previous data
Restored to factory default settings: By Factory data
Restored to previous settings: By Previous data
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
8.6
8-16
<8. YH8000 HMI Unit>
Configuration Screen
The configuration screen is used to configure the TDLS8000 and YH8000.
Tapping
at the bottom of the screen shows the setup target selection screen.
Tag name
Alarm icon
HMI selection
Figure 8.18
Setup target selection screen (when a single TDLS8000 is connected)
Tag name
Shows the tag names assigned to each TDLS8000. Serial numbers are displayed if tag
names are not assigned.
Alarm icon
If an alarm is occurring on the TDLS8000, an alarm icon is displayed. If both a warning
are occurring, the fault icon
is displayed.
and a fault
HMI selection
8.6.1
To switch to the YH8000 configuration screen, select the HMI option.
TDLS8000 configuration screen
After selecting a TDLS8000 on the setup target selection screen, tap OK. A password input
screen will appear.
Figure 8.19
Password input screen
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
8-17
<8. YH8000 HMI Unit>
Enter the password on the password input screen, and tap
configuration screen.
to display the TDLS8000
Execution menu
Configuration menu
Figure 8.20
TDLS8000 configuration screen
Execution menu
The execution operation screens for calibration, validation, loop check, and the like are
under this menu.
Configuration menu
Various settings are under this menu.
The TDLS8000 configuration screen is arranged a menu tree structure.
• For the menu structure, see “Appendix 4 YH8000 Menu Tree”.
• For details on various settings, see “6.
Configuration”.
• For details on calibration and validation, see “9.
Inspection and Maintenance”.
• For details on loop check, see “5.4 Loop Check (Simulation output)”.
NOTE
TDLS8000 status
When you switch to the TDLS8000 configuration screen, the TDLS8000 status is set to “in
maintenance.” While in maintenance, the TDLS8000 cannot be configured from other YH8000s
or HART. Tap any of the buttons at the bottom of the screen to exit from the TDLS8000
configuration screen. When you exit from the configuration screen, “in maintenance” is cleared.
“In maintenance” is also cleared when the connection to the TDLS8000 is disconnected.
NOTE
Password
The factory default password for switching to the TDLS8000 configuration screen is “1234.” You
can change the password on the following configuration screen.
“
>>[Analyzer]>>Configuration>>System>>Password”
For the password, you can set a character string consisting of one to eight characters of your
choice.
Keep the password in a safe place.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
8-18
<8. YH8000 HMI Unit>
NOTE
If you forget the password
If you forget the password, contact your nearest Yokogawa representative.
If you can use a HART configuration tool, you can initialize the user information to reset the
password to the factory default setting. However, other settings will also be initialized. For details,
see “6.11.1
Initialization Procedure”.
8.6.2
YH8000 Configuration Screen
After selecting an HMI on the setup target selection screen, tap OK to display a YH8000
configuration screen. If there is no connected TDLS8000, only the YH8000 configuration screen
will be displayed.
Analyzer connection menu
Screen configuration menu
IP address configuration menu
Language configuration menu
HMI information screen
Figure 8.21
YH8000 (HMI) configuration screen
Analyzer connection menu
Switches to the TDLS8000 selection screen. For details, see “5.2.3 Connecting to the
TDLS8000”.
Screen configuration menu
Switches to the configuration screen for setting the home screen display style, meter range,
and backlight. For the display styles, see “8.2.2
Selecting the Style”. For the meter
range, see “8.2.3 Setting the Meter Range”. For backlight settings, see “8.6.3
Setting
the YH8000 Backlight”.
IP address configuration menu
Switches to the YH8000 IP address configuration screen. For details, see “5.2.2
the IP Address”.
Setting
Language configuration menu
Shows the language displayed on the YH8000.
HMI information screen
Shows the YH8000 IP address, MAC address, and software version.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
8.6.3
<8. YH8000 HMI Unit>
8-19
Setting the YH8000 Backlight
This section explains how to set the brightness of the YH8000 backlight and the auto-off function.
Configuration menu path:
“
>>HMI>>Display Setting>>Backlight”
Auto Off
Brightness
Figure 8.22
YH8000 backlight configuration screen
Brightness
You can set the backlight brightness using 11 levels. By factory default, the brightness is set
to the highest level.
Auto Off
The backlight automatically turns off after a certain time of no YH8000 touch panel activity.
You can set the auto-off time to any of the options in the following table.
Backlight auto-off time
Available option
10 min
30 min
60 min
Always On
Description
Turns off after 10 minutes of no activity
Turns off after 30 minutes of no activity
Turns off after 60 minutes of no activity
Disables the auto-off function (always on)
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
8.7
8-20
<8. YH8000 HMI Unit>
When Multiple TDLS8000s Are Connected
When multiple TDLS8000s are connected to the YH8000, a portion of the display and operation
is different. This section explains the difference.
8.7.1
Overall Display
When multiple TDLS8000s are connected, the information of all the TDLS8000s can be
collectively displayed on the home screen. This is called the overall display.
Date and time
Display switch tab
Analyzer number
Pressure icon
Temperature icon
Alarm
Icon
Figure 8.23
Overall display (when three TDLS8000s are connected)
Date and time
On the overall display, the time of the TDLS8000 with the smallest analyzer number is
shown.
Analyzer number
Shows analyzer numbers 1 to 4.
Analyzer numbers are assigned using the YH8000 analyzer connection settings. For details,
see “5.2 Connecting to the YH8000”.
Alarm icon
If an alarm is occurring on the TDLS8000, an alarm icon is displayed. If both a warning
are occurring, the fault icon
is displayed.
and a fault
Display switch tab
Selects the TDLS8000 to be displayed. Selecting
shows the overall display.
Temperature icon and pressure icon
Shows the process temperature and process pressure mode settings.
: Fixed value
: Non-fixed value
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
8.7.2
8-21
<8. YH8000 HMI Unit>
TDLS8000 Selection Screen
When multiple TDLS8000s are connected, a TDLS8000 selection screen appears when
switching to the
alarm screen or
information screen. Selecting the target TDLS8000 and
touching OK will display the alarm screen or information screen of the selected TDLS8000.
Analyzer number
Tag name
Alarm icon
Figure 8.24
TDLS8000 selection screen (when three TDLS8000 are connected)
Analyzer number
Shows analyzer numbers 1 to 4.
Tag name
Shows the tag names assigned to each TDLS8000. Serial numbers are displayed if tag
names are not assigned.
Alarm icon
If an alarm is occurring on the TDLS8000, an alarm icon is displayed. If both a warning
are occurring, the fault icon
is displayed.
and a fault
When multiple TDLS8000s are connected, the connected TDLS8000s are shown on the setup
configuration screen. Selecting the target
target selection screen when switching to the
TDLS8000 and touching OK will display the password input screen of the selected TDLS8000.
Analyzer number
Tag name
Alarm icon
HMI selection
Figure 8.25
Setup target selection screen (when three TDLS8000 are connected)
Analyzer number
Shows analyzer numbers 1 to 4.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
8-22
<8. YH8000 HMI Unit>
Tag name
Shows the tag names assigned to each TDLS8000. Serial numbers are displayed if tag
names are not assigned.
Alarm icon
If an alarm is occurring on the TDLS8000, an alarm icon is displayed. If both a warning
are occurring, the fault icon
is displayed.
and a fault
HMI selection
To switch to the YH8000 configuration screen, select the HMI option.
8.7.3
Setting the Date and Time on the TDLS8000
If multiple TDLS8000s are connected, the following function can be used when setting the
TDLS8000 clocks.
• Change all analyzer
You can collectively set the clocks on all connected TDLS8000s.
• Time synchronization
You can synchronize the clocks of other connected TDLS8000s with the clock of the
TDLS8000 that you are currently configuring.
Configuration menu path:
“
>>[Analyzer]>>Configuration>>System>>Date&Time>>Operation”
NOTE
The collective setting of the clocks and synchronization of the clocks do not completely
synchronize the TDLS8000 clocks. Even when these functions are used, the clocks may be off
by a few seconds. The clocks may drift even further as time passes.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9.
9-1
<9. Inspection and Maintenance>
Inspection and Maintenance
CAUTION
If you need to remove the TDLS8000 from the process flange for inspection or maintenance, be
sure to turn off the power beforehand.
Work performed by an unqualified engineer can cause injury or severe damage to instruments.
Not following the warnings in this manual can also cause injury or severe damage to instruments.
Make sure that maintenance is carried out by a qualified engineer. A qualified engineer is an
engineer who:
• Is knowledgeable about the safe handling of process analysis instruments (or general
automation technology) and has read and understood the content of this manual.
• Has received instructions on how to start and configure instruments and has read and
understood the content of this manual.
This chapter explains inspection and maintenance to retain the measurement performance of the
TDLS8000.
There are no operations that need to be performed regularly on the TDLS8000.
When the laser beam transmission decreases, clean the process window.
9.1
Maintaining the Laser Beam and
Transmission
Transmission is a value determined by the magnitude of laser power that reaches the photo
detector in the sensor control unit (SCU) after the laser beam emitted from the laser element
of the LU passes through the gas to be measured. Transmission is used to verify aging after
the optical axis is adjusted. As startup, adjust the optical axis correctly, perform transmission
calibration, and set the transmission value to 100%.
By checking the variation in the transmission after startup, you can determine the state of the
region that the laser beam travels through, the degree of optical axis misalignment between
the LU and SCU, and the states of the laser beam emitting and receiving components. This
information is important for maintenance and troubleshooting.
Generally, the transmission degrades due to the conditions shown below.
To maintain normal TDLS8000 operation, perform the required inspection and maintenance
to prevent the transmission from degrading. To maximize the TDLS8000 performance, it is
particularly important to optimally adjust the optical axis and keep the process window clean.
l Stained process window
If the process window is stained, the laser beam will be blocked causing the amount of laser
beam reaching the photo detector to decrease. Perform process window purge continuously to
prevent stains from adhering to the process window. See “3.4 Piping”.
l Particles (dust and the like) in the process
Particles in the process gas block the laser beam causing the amount of laser beam reaching
the photo detector to decrease. For measurement locations (applications) with high level of
particles in the process, take measures such as installing insertion tubes as described in “3.1.3
Attaching the TDLS8000 to the Process Flange” in addition to process window purging.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9-2
<9. Inspection and Maintenance>
l LU and SCU optical axis misalignment
With a duct or flue with thin and easily bent walls, optical axis misalignment will occur between
the LU and SCU after the TDLS8000 is installed, degrading transmission. Refer to “3.1.2
Constructing Process Flanges”, and check the construction of the installation location. Also, in
applications that measure the gas concentration in high temperature gas, bending caused by
heat during processing may cause the optical axis to become misaligned. If this happens, refer to
“3.3 Optical Axis Adjustment”, and readjust the optical axis to the correct alignment.
l Optical axis adjustment of the LU and SCU after removing alignment flanges
If you remove alignment flanges to perform inspection and maintenance, be sure to perform
optical axis adjustment according to “3. Installation, Wiring, Optical Axis Adjustment, and Piping”
and “3.3
Optical Axis Adjustment” when you reinstall the LU and SCU in the measurement
location.
9.1.1
Transmission Calibration
After reinstalling the TDLS8000 in the measurement location and performing optical axis
adjustment according to “3.3 Optical Axis Adjustment”, perform transmission calibration.
NOTE
Optimum transmission may not be attainable if the optical axis is not adjusted correctly. Refer to
“3.3 Optical Axis Adjustment”, and perform optical axis adjustment correctly.
Execution menu path:
[HART] “Diagnosis/Service>>Transmission adjust>>Transmission adjust”
[YH8000] “ >>Execution>>Transmission Adjustment>>Measurement>>Transmission
Adjustment”
9.1.2
Process Window Cleaning
Under normal operation, if process window purge is performed correctly, the surface of the
process window rarely become stained. However, the surface of the process window may
become stained or clouded under the following conditions.
• If the gas to be measured including dust or stain makes contact with the process window
due to insufficient process window purge flow rate or purge gas pressure.
• If the surface of the process window condenses when the process window is hot
• If gas that would cause quality deterioration in the process window (e.g., hydrogen fluoride
on BK-7) makes contact with the window
• If particles, oil, and the like from the purge gas facility adheres to the surface of the process
window
If the surface of the process window is stained, remove and clean the process window according
to the following procedure.
CAUTION
Before removing the process window, check that the process is completely stopped or that the
process is isolated from open air and no process gas will be discharged.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9-3
<9. Inspection and Maintenance>
CAUTION
Be careful in handling the process window as it is made of optical glass that is easily damaged.
(1) Turn off the TDLS8000.
(2) Remove the TDLS8000 from the process.
(If necessary, separate it completely from the process such as by using a process isolation
valve.)
(3) Check the stained area of the process window, and remove the relevant process window.
Process windows are installed in alignment flanges, process isolation flanges, flow cells,
and the like.
(4) Loosen the four M4 screws with hexagonal holes on the process window holder installed in
the alignment flange, and remove the process window.
For details on how to install and remove the process window, see Figures 9.1 to 9.3.
(5) Using clean and dry instrumental air or nitrogen gas, blow off the particles from the surface
of the process window.
(6) Using warm water and low irritative cleansing agent, gently wipe the surface of the process
window with a soft cloth, being careful not to scratch it. Then, if necessary, clean with alcohol
(e.g., isopropyl alcohol).
(7) Blow clean and dry instrumental air or nitrogen gas again on the surface to dry it.
(8) Thoroughly examine the surface of the process window from various angles, and check that
the stain has been sufficiently removed and that the process window is ready for use.
NOTE
If you cannot remove the stain from the surface of the process window, replace with a new one.
If the surface of the process window is corroded, it may have been contaminated with corrosive
gas such as hydrogen fluoride. In such a case, replace the process window with a new one.
(8) Install the cleaned process window (or a new one).
Pay attention to the orientation of the process window. Install it in the same orientation as
before.
Tighten the screws evenly.
(9) After installing the process window, install the TDLS8000 for use.
If you removed an alignment flange from the process in order to clean the process window
of a process isolation flange or if you removed an alignment flange from a flow cell to clean
the process window of the flow cell, after installing the alignment flange, readjust the optical
axis before use.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9-4
<9. Inspection and Maintenance>
Alignment flange
Process window
O-ring
Hole with a hexagonal hole
Alignment flange
Surface to be cleaned
Figure 9.1
Removing and installing the process window of an alignment flange
Process isolation flange
Process window
O-ring
Washer
Hole with a hexagonal hole
Process isolation flange
Surface to be cleaned
Figure 9.2
Removing and installing the process window of a process isolation flange
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9-5
<9. Inspection and Maintenance>
Hole with a hexagonal hole
Process window
Flow cell
O-ring
Washer
Surface to be cleaned
Flow cell
Figure 9.3
9.1.3
Removing and installing the process window of a flow cell
Insertion Tube Cleaning
Since insertion tubes are used in a process where there is a lot of dust, long-term use can cause
dust or the like to accumulate inside the tubes at the tip. Regularly clean insertion tubes.
The cleaning procedure is as follows.
(1) Turn off the TDLS8000.
(2) Turn off the process window purge gas.
(3) Remove the TDLS8000.
(4) Remove the process interface (alignment flange or process isolation flange (if installed)).
(5) Remove the insertion tube.
(6) Remove the dust or the like that has accumulated inside the insertion tube at the tip. If
necessary, clean with a cleansing agent or the like.
(7) Install the insertion tube.
(8) Install the process interface (alignment flange or process isolation flange (if installed)).
(9) Install the TDLS8000, and turn on the power.
(10) Adjust the optical axis.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9.2
9-6
<9. Inspection and Maintenance>
Online Validation
Online validation is performed by purging a validation cell with check gas of known concentration
while measuring the concentration of the measured gas under stable measured gas
concentration conditions. The conditions that are controlled (or known) when purging with check
gas are as follows.
• Pressure of the purge check gas
• Temperature of the purge check gas
• Length of the validation cell purged with check gas
• Concentration of the purge check gas
The basic procedure is shown below.
• Set known validation parameters.
• Purge the validation cell with check gas of known concentration.
=> The result of “Process concentration” + “additional concentration from the check gas” is
recorded in the TDLS8000.
• Purge the validation cell again with the original purge gas (typically nitrogen gas).
=> “Process concentration” is recorded in the TDLS8000.
=> The expected value for the “additional concentration from the check gas” is calculated
from the known parameters.
=> The expected value and the actual value are compared and validated (pass or fail).
NOTE
Perform online validation when the process is stable. Validation cannot be performed accurately
when the concentration, temperature, or pressure of the process gas is fluctuating greatly.
NOTE
Validation is a procedure to check whether the TDLS8000 is operating properly. If there is a
reading error as a result of validation, check that there are no gas leaks from the process. If there
is no gas leak, perform calibration.
9.2.1 Preparation
In online validation, a validation cell is purged with check gas. There are two check gas piping
methods.
• When there is no process gas switching
Up to two check gases are connected for online validation 1 and online validation 2 (Figure
9.4).
• When there is process gas switching
A single check gas is connected for online validation 2. The remaining stream is used to
switch between two process gases (Figure 9.5).
NOTE
If you want to switch the check gas stream through automatic valve control using the TDLS8000
SV terminal, you need to set the TDLS8000 valve usage according to the piping method as
shown in the following table. For details on valve usage, see “6.8.2
Valve Usage Setting”.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9-7
<9. Inspection and Maintenance>
Piping
Valve usage setting
When there is no process gas switching Cal/Val
When there is process gas switching
2 Streams & Cal/Val
Online validation piping diagram is shown below.
l When there is no process gas switching
Connect up to two types of online validation check gas. Execute online validation 1 or 2 to
validate using the respective check gas. If you want to control valves automatically through the
SV terminal, you need to set the valve usage to Cal/Val.
Laser unit (LU)
Sensor control unit (SCU)
Measured gas
Purge gas
Process gas
(Stream1)
Valve-2
(SV-2)
Online Validation 2
(Stream3)
Figure 9.4
Valve-1
(SV-1)
Online Validation 1
(Stream2)
Online validation piping diagram for when there is no process gas switching
l When there is process gas switching
Connect two types of process gas and a single type of online validation gas. The process gas
concentration affects the results of online validation. As such, switch to the process gas to be
validated, and then start the validation.
If you want to control valves automatically through the SV terminal, you need to set the valve
usage to 2 Streams & Cal/Val. Only online validation 2 can be executed.
NOTE
To automatically control valves in this arrangement, set the current stream to the same stream as
the initial stream before starting validation. For example, if the initial stream is Stream1, switch to
Stream1, and then start validation. Do not switch to Stream2 until the validation is finished.
The piping diagram for when the valve usage is 2 Streams & Cal/Val is shown in Figure 9.5.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9-8
<9. Inspection and Maintenance>
Laser unit (LU)
Sensor control unit (SCU)
Flow cell
Purge gas
Process gas 1
(Stream1)
Purge gas
Valve-2
(SV-2)
Valve-1
(SV-1)
Process gas 2
(Stream2)
Online Validation
Figure 9.5
9.2.2
Online validation piping diagram for when there is process gas switching
Configuration
Online validation configuration menu:
[HART] “Detailed setup>>Validation>>Online Validation #”
[YH8000] “
>>Configuration>>Validation>>Online Validation #”
The setup parameters required to manually execute online validation are indicated for each of the
above submenus (tabs on the YH8000). Here, online validation 1 will be used as an example.
l Parameter
Parameter name
(HART)
Onval1 gas type
Onval1 gas conc
Onval1 temp mode
Onval1 temp fix val
Onval1 act amb
ofst
Onval1 pres fix val
Onval1 OPL fix val
*1:
Parameter name
(YH8000)
Gas type
Concentration
Temperature
Fixed Value
Offset Value
Pressure
OPL
Description
Selects the type of validation 1 check gas (two-gas measurement only)
Enters the concentration of the validation 1 check gas
Selects the temperature mode for validation 1 execution
Enters the temperature for when Onval1 temp mode is set to Fixed
Enters the temperature offset for when Onval1 temp mode is set to Active
ambient
Enters the pressure value for validation 1 execution
Enters the process optical path length of the region purged with validation
1 check gas (*1)
The process optical path length of online validation is determined by the TDLS8000 model and code as shown in the following
table.
Model and code
Optical path length [mm]
Other than -LA
130.6
-X1
-LA
235.6
Other than -LA
130.6
-G1 -X2 -LA
235.6
-G2
Other
than
-LA
92.6
-D2 -C3
-LA
225.6
TDLS8000 -C2
-S2
Other than -LA
92.6
-E2 -C4
-LA
225.6
-J2
-A1
92.6
-H1
132.5
-H3
132.5
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9-9
<9. Inspection and Maintenance>
l Valve
Parameter name
Description
(HART)
Onval1 auto vlv man Selects whether to enable automatic valve control through the SV terminal during
manual validation 1 execution.
Onval1 gas purg time For automatic execution. Set these when performing automatic execution (see section
Onval1 nml purg time 9.8.2).
l Auto time
For automatic execution. Set this when performing automatic execution (see “9.8.2
Configuration”).
l Reading mode
For automatic execution.
Parameter name
(HART)
Onval1 read mode
Description
Selects the concentration reading setting for automatic validation.
If “Process+Validation” is selected, the reading shows the sum of the process and
validation cell concentrations.
If “Validation only” is selected, the reading shows only the validation cell concentration.
Onval1 output factor Scaling coefficient for the concentration reading during automatic validation. If
Reading mode is set to Validation only, the reading will be the product of the calculated
concentration and the scaling coefficient.
9.2.3
Execution
Before starting online validation, check that the piping and online validation settings are correct.
Here, online validation 1 for O2 will be used as an example.
Execution menu path:
[HART] “Diagnosis/Service>>Validation>>Manual>>Manual online val 1”
[YH8000] “
>>Execution>>Validation>>Manual>>Online Validation 1”
n HART execution screen
(1) Starting an online validation
If automatic valve control is disabled, the following screen will appear.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<9. Inspection and Maintenance>
9-10
If automatic valve control is enabled, a message stating that the valves will be automatically
controlled will appear as follows. In this case, you do not need to manually control the valves
during validation.
In either screen, check the type of gas to use as check gas, and tap OK.
On the next screen, tap ENTER. If automatic valve control is enabled, the stream will be
switched, and the validation cell will be purged with check gas.
(2) Purging with the check gas
If automatic valve control is disabled, manually control the valves to purge the validation cell
with the check gas. By referring to the “stdev” value, which indicates the standard deviation
of concentration, check that the concentration is stable over a sufficient length of time (5
minutes as a guideline, at least 1 minute) with the validation cell filled with check gas. When
stability is confirmed, tap OK.
If automatic valve control is enabled, the stream will be switched automatically, and check
gas will be discharged from the validation cell.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<9. Inspection and Maintenance>
9-11
(3) Discharging the check gas
If automatic valve control is disabled, manually control the valves to purge the validation
cell with the analyzer internal purge gas (nitrogen gas) that is used normally during process
measurement. By referring to the “stdev” value, which indicates the standard deviation of
concentration, check that the concentration is stable, and then tap OK. The validation result
will be displayed.
(4) Checking the online validation result
The online validation result is displayed, and online validation ends. If successful, “PASSED”
appears. Otherwise, “FAILED” appears.
Expected reading: Gas concentration (expected value) obtained by adding the
check gas
Actual reading:
The actual value
Tap OK to return to the menu.
NOTE
If the validation fails, the following warning will occur. For the corrective action, see “10.2 Warning
Display and Handling”.
Alarm number
Alarm name
15
Validation Error
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<9. Inspection and Maintenance>
9-12
n YH8000 Execution Screen
(1) Starting an online validation
Touch Start to begin online validation. If automatic valve control is enabled, the stream will
be switched, the validation cell will be purged with check gas.
(2) Purging with the check gas
If automatic valve control is disabled, manually control the valves to purge the validation cell
with the check gas. Check that the concentration is stable over a sufficient length of time (5
minutes as a guideline, at least 1 minute) with the validation cell filled with the check gas.
When stability is confirmed, tap Next. If automatic valve control is enabled, the stream will
be switched automatically, and check gas will be discharged from the validation cell.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<9. Inspection and Maintenance>
9-13
(3) Discharging the check gas
If automatic valve control is disabled, manually control the valves to purge the validation
cell with the analyzer internal purge gas (nitrogen gas) that is used during process
measurement. Check that the concentration is stable, and then touch Execute. The
validation result will be displayed.
(4) Checking the online validation result
The online validation result is displayed, and online validation ends. If successful, “PASSED”
appears. Otherwise, “FAILED” appears.
Expected Reading: Gas concentration (expected value) obtained by adding the
check gas
Actual Reading:
The actual value
Tap OK to return to the configuration menu.
NOTE
If the validation fails, the following warning will occur. For the corrective action, see “10.2 Warning
Display and Handling”.
Alarm number
Alarm name
15
Validation Error
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<9. Inspection and Maintenance>
9.2.4
9-14
Time Chart
The valve operation during manual online validation execution and the timing when the AO/DO
output switches to Cal/Val mode are shown below. In Cal/Val mode, it is possible to hold the AO
output or specify other settings. For the AO/DO output settings during Cal/Val mode, see “6.4.2
Output Hold” and “6.5.1 DO Contact (DO-1)”.
In Figure 9.6, Valve1 and Valve2 are switched manually by following the instructions on the
operation screen. If automatic valve control is enabled, there is no need for the operator to switch
the valves manually.
Time
Operator
HART/HMI screen Valve1 Valve2
[Screen operation] Start online validation 1
OFF
OFF
ON
OFF
AO/DO
mode
Normal output
[Manual valve operation] Purge with check gas
(Wait for the gas concentration to stabilize.)
Check gas purging
[Screen operation] Proceed to the next step
Cal/Val
[Manual valve operation] Discharge check gas
(Wait for the gas concentration to stabilize.)
Check gas
discharging
OFF
OFF
[Screen operation] Touch Execute
Normal output
Validation result
Figure 9.6
Time
Valve and AO/DO output for manual online validation 1
Operator
HART/HMI screen Valve1 Valve2
[Screen operation] Start online validation 2
OFF
OFF
ON
ON
AO/DO
mode
Normal output
[Manual valve operation] Purge with check gas
(Wait for the gas concentration to stabilize.)
Check gas purging
[Screen operation] Proceed to the next step
Cal/Val
[Manual valve operation] Discharge check gas
(Wait for the gas concentration to stabilize.)
Check gas
discharging
OFF
OFF
[Screen operation] Touch Execute
Normal output
Validation result
Figure 9.7
Valve and AO/DO output for manual online validation 2
If the initial stream is set to Stream1 in a piping arrangement with process gas switching, the
valve state transition is different as shown below. For details, see section 9.2.1.
Operator
[Screen operation] Start online validation 2
[Manual valve operation] Purge with check gas
[Manual valve operation] Discharge check gas
9.3
Valve1 Valve2
OFF
OFF
OFF
OFF
ON
OFF
Mounting on a Calibration Cell
Before performing offline validation, zero calibration, or span calibration, you usually need
to remove the LU and SCU from the process interface and mount them on a calibration cell.
However, if the TDLS8000 is mounted on a flow cell, the gas flowing through the flow cell can be
switched from the process gas making it possible to perform offline validation, zero calibration,
and span calibration without removing the TDLS8000 from the flow cell.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9-15
<9. Inspection and Maintenance>
Calibration Cell
Figure 9.8
Connection example for offline work
9.3.1 Preparation
If you want to perform calibration or offline validation, prepare the following tools, instruments,
nitrogen gas, and gas for offline work (check gas, zero calibration gas (nitrogen gas), span
calibration gas).
No.
Tool or instrument
1 Calibration cell
2 YH8000 or HART configuration
tool
3 Cable between the LU and SCU
4 24 V DC power cable
5 Valve drive cable
1
1
As required (1 or 2)
6
7
8
9
10
11
12
13
14
15
16
1
1
Several meters
As required
1
2
As required
1
1
As required
As required
YH8000 connection cable
24 V DC power supply
1/4 inch piping
1/4 inch ferrule set
Pressure regulator
1/4 inch pipe plug
Three-way valve
Thermometer
Pressure meter
Coupling
Nitrogen gas
Quantity
1
1
17 Span calibration gas
18 Check gas
19 Flowmeter
As required
As required
3
20 Needle valve
3
Remarks
For executing offline calibration and validation
For controlling valves automatically through the
SV terminal
When using the YH8000
For TDLS8000 purging
For zero calibration
For span calibration
For offline validation
For TDLS8000 purging
For zero calibration
For span calibration and offline validation
For flow rate adjustment
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9.3.2
9-16
<9. Inspection and Maintenance>
Preparation Procedure
Perform the following procedure. Purge pipe connection is different for offline validation, zero
calibration, and span calibration. For details, see sections 9.4, 9.5, and 9.6.
CAUTION
• Do not apply physical shock to the TDLS8000 when relocating the TDLS8000 to a
calibration cell and when returning it to the process. Doing so can cause a malfunction.
• During calibration work, do not remove the LU or SCU while the power is on.
• If the process gas is positive pressure, shut off the TDLS8000 from the process, stop the
process window purge, and prevent excessive pressure from being applied to the process
window.
(1) Recording the settings
Before removing the TDLS8000, check the following settings and the process conditions.
These will be used when returning the TDLS8000 to the process.
• Process optical path length
• Process pressure (record only when the input mode is set to Fixed)
• Process temperature (record only when the input mode is set to Fixed)
• Transmission
(2) Turning the TDLS8000 off
Turn the TDLS8000 off.
(3) Removing the TDLS8000 from the process
(a) Stopping the purge gas
Stop the nitrogen gas (or instrumental air) for process window purging and analyzer internal
purging.
Prevent pressure from being applied between the inside of the TDLS8000 and the ball valve or
between process windows (attached to alignment flanges or process isolation flanges) and the
ball valve.
(b) Shutting off the TDLS8000 side from the process side (using a ball valve)
Check that there is no residual internal pressure, and then shut off the TDLS8000 from the
process. (Close the process ball valve or the like.) If there is no component for shutting off the
TDLS8000 side from the process side, check that the process is completely stopped before
removing the TDLS8000. If there is no component for shutting off the TDLS8000 side from the
process side and the process is running, do not remove the TDLS8000 as the process gas may
blow out.
(c) Removing piping
Remove the pipes from the TDLS8000. (To make reinstallation easier after offline validation or
calibration, we recommend that you mark the pipes.)
Apply vinyl tape or other protector to the TDLS8000 ports and pipe ferrule areas.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9-17
<9. Inspection and Maintenance>
(d) Removing wiring
Remove wiring.
Be careful not to short the wires. Insulate and protect the removed wires with vinyl tape or the
like, and bundle them together, making sure not to strain the cables. (For details on wiring, see
“3.2 Wiring”.)
NOTE
To make reinstallation easier after the work is complete, we recommend that you mark the wires.
(e) Removing the TDLS8000 (LU and SCU)
Check that piping and wiring have been completely removed, and then remove the TDLS8000.
The method to remove the LU and SCU is the same. Remove them one at a time.
If a YH8000 is installed, remove the entire YH8000 with its mounting bracket before removing the
SCU.
(1) Using a hex wrench (5 mm), remove only the upper right screw of the quick connector (see
Figure 9.9).
(2) Loosen the other screws (upper left, lower left, and lower right).
(3) Slowly turn the TDLS8000 counterclockwise to remove it from the alignment flange.
For details on how to remove the YH8000, see “4. YH8000 Installation”.
Hole with a hexagonal hole
Quick connector
Turn counterclockwise
Process flange
Alignment flange
Figure 9.9
Removing from the process
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9-18
<9. Inspection and Maintenance>
(4) Mounting on a Calibration Cell
There are three screws with hexagonal holes at the upper left, lower left, and lower right on each
of the calibration cell interface to the TDLS8000. Check that these screws are loosened enough
(at least 8 mm) so that the LU or SCU can be mounted.
Hole with a hexagonal hole
8 mm or more
Figure 9.10
Gap in the TDLS8000 mounting screws with hexagonal holes
(a) Mounting the LU and SCU
Mount the LU or SCU to one end of the calibration cell. While making sure that the O-ring does
not come loose or it is not damaged during the mounting process, insert the quick connector into
the screw area, and rotate the LU or SCU clockwise.
While holding the LU or SCU perpendicular to the mounting surface, tighten the four screws,
including the upper right screw, evenly.
Finally, mount the YH8000.
(b) Wiring
Connect the following cables.
• Inter-unit cable
• Power cable
• Valve drive cable (when necessary)
For details on wiring, see “3.2 Wiring”.
(c) Piping
When the TDLS8000 is mounted on a calibration cell, the resulting unit is divided into three
sections as shown in Figure 9.11. For Explosionproof/Flameproof Type, devided into five
sections.
An appropriate gas must be fed into each section. Connect the pipes according to the offline work
outlined in sections 9.3 to 9.5.
Stainless tubes or Teflon tubes are used for piping.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9-19
<9. Inspection and Maintenance>
LU
Analyzer internal purge
discharge
Nitrogen gas*2
Sealed
or
Exhaust
Nitrogen gas
*1 *3
Figure 9.11
Calibration cell
SCU
Zero gas (nitrogen gas) Internal calibration cell
or span calibration gas
purge discharge
or check gas*
Analyzer internal purge
discharge
Nitrogen gas*2
*1: Max. output: 0.02 MPa
Normal flow rate: < 2 L/min
*2: For Explosionproof/Flameproof type
*3: For Explosionproof/Flameproof type,
flow rate is 150 ml/min
Sealed
or
Exhaust
Nitrogen gas*1 *3
Purge gas piping diagram
When piping is complete, check for leaks such as by using Snoop. Use nitrogen gas for this
purpose. For Explosionproof/Flameproof Type, at 150 mL/min, no more than 0.04 MPa.
(5) Feeding purge gas
When wiring and piping are complete, feed the gas.
Feed the appropriate gas for offline work at a flow rate no more than 2 L/min at a pressure no
more than 0.02 MPa.
(6) Turning the power on
Turn the power on. Check that the TDLS8000 starts normally.
9.3.3
Performing Calibration and Offline Validation
Refer to sections 9.4, 9.5, and 9.6, and perform offline work.
For each kind of offline work, the settings may be different. Change the settings as necessary.
9.3.4
Returning the TDLS8000 to the Process
When offline work is complete, return the TDLS8000 to the process. Follow the procedure below.
NOTE
Check that the process is stopped before proceeding with this procedure.
(1) Switching the span calibration gas or check gas
If hazardous gas is flowing (e.g., CO gas), switch it to nitrogen gas. Wait for the gas to be
completely replaced before proceeding.
(2) Stopping the gas
On the TDLS8000 display, check that the calibration gas concentration has dropped to zero,
and then stop all purge gases.
(3) Removing piping
Check that there is no residual internal pressure, and then remove the piping.
(4) Turning the TDLS8000 off
Check the above items, and then turn the TDLS8000 off.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<9. Inspection and Maintenance>
9-20
(5) Removing wiring
Remove wiring according to the procedure of section 9.3.2 (3) (d).
(6) Removing the TDLS8000 from the calibration cell
Remove the TDLS8000 according to the procedure of section 9.3.2 (3) (e).
(7) Installing the TDLS8000 in the process
Install the TDLS8000 by reversing the procedure for removing it.
(a) Mounting the LU and SCU
Attach the SCU or LU to the alignment flange. While making sure that the O-ring
does not come loose or it is not damaged during the mounting process, insert the
TDLS8000 quick connector into the alignment flange, and rotate the TDLS8000
clockwise.
While holding the LU or SCU perpendicular to the mounting surface, tighten the four
screws, including the upper right screw, evenly.
(b) Wiring
Connect the following cables.
•
Cable between the LU and SCU
•
Power cable
•
Valve drive cable (when necessary)
•
AI/AO/DO/DI cable (when necessary)
For details on wiring, see “3.2
Wiring”.
(c) Piping
Refer to “3.4 Piping”, and connect the pipes to restore the TDLS8000 to its original
condition before it was removed.
(8) Feeding purge gas
When wiring and piping are complete, feed the purge gas.
Since the internal pressure from the process window purge gas may become high when
process isolation valve (ball valve) is used, open the valve immediately after starting to feed
the process window purge gas.
(9) Turning the power on
Turn the power on. Check that the TDLS8000 starts normally.
(10) Optical Axis Adjustment
Refer to “3.3
Optical Axis Adjustment”, and adjust the optical axis.
(11) Checking the settings
Refer to the settings that you recorded before removing the TDLS8000, and reset them if
necessary.
a) Process optical path length
b) Process pressure
c) Process temperature
d) Transmission
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9.4
9-21
<9. Inspection and Maintenance>
Offline Validation
Offline validation is a function used to verify the validity of gas concentration measurements. For
the validation process, the TDLS8000 is separated from the measurement process, and a known
check gas if fed through a calibration cell or flow cell.
Before performing a validation, you need to enter the following information in the TDLS8000.
• Pressure of the purge check gas
• Temperature of the purge check gas
• Length of the cell to be purged with check gas
• Concentration of the purge check gas
The basic procedure is shown below.
• Set known validation parameters.
• Purge the calibration cell or flow cell with check gas of known concentration.
=> The check gas concentration reading will be recorded.
=> The expected value for the “check gas” is calculated from the known parameters.
=> The expected value and the actual value are compared and validated (pass or fail).
NOTE
Validation is a procedure to check whether the TDLS8000 is operating properly. If there is an
error in reading, calibrate.
Normally in an in-situ installation, the LU and SCU are removed from the process and mounted
to a calibration cell before performing offline validation. In a flow cell or bypass installation, offline
validation can be performed without removing the LU and SCU by switching the stream from the
process gas to check gas.
9.4.1 Preparation
Follow the instructions in “9.3 Mounting on a Calibration Cell”. There are two piping methods
for offline validation.
l Offline validation exclusive
Connect both offline validation 1 and 2 check gases. Validation 1 and 2 can be executed.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9-22
<9. Inspection and Maintenance>
Laser unit (LU)
Sensor control unit (SCU)
Flow cell
Purge gas
Purge gas
Process gas
(Stream1)
Valve-2
(SV-2)
Online Validation 2
(Stream3)
Valve-1
(SV-1)
Online Validation 1
(Stream2)
In this piping arrangement, automatic offline validation 1 and 2 can be executed consecutively.
There are two methods to start a consecutive execution. One is by using HART, YH8000,
Modbus, or digital input (semi-automatic execution), and the other is to start at a specified
time (automatic execution). For details on automatic and semi-automatic execution, see “9.8
Automatic and Semi-automatic Execution of Validation and Calibration”.
For details on consecutive execution, see “9.8.5 Consecutive Automatic Execution”.
l Offline validation check gas and two types of process gas
Offline validation check gas is connected. The remaining stream is used to switch between two
process gases.
Laser unit (LU)
Sensor control unit (SCU)
Flow cell
Purge gas
Purge gas
Process gas 1
(Stream1)
Valve-2
(SV-2)
Valve-1
(SV-1)
Process gas 2
(Stream2)
Online Validation
(Stream3)
If you want to switch the check gas stream through automatic valve control using the TDLS8000
SV terminal, you need to set the TDLS8000 valve usage according to the piping method as
shown in the following table. Note that only online validation 2 can be executed. For details on
valve usage, see “6.8.2 Valve Usage Setting”.
Piping
Offline validation exclusive
Offline validation 2 check gas and two
types of process gas
Valve usage setting
Cal/Val
2 Streams & Cal/Val
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9-23
<9. Inspection and Maintenance>
9.4.2
Configuration
Offline validation configuration menu:
[HART] “Detailed setup>>Validation>>Offline validation #”
[YH8000] “
>>Configuration>Validation>>Offline Validation #”
The setup parameters required to manually execute offline validation are indicated for each of the
above submenus (tabs on the YH8000). Here, offline validation 1 will be used as an example.
l Parameter
Parameter name
(HART)
Offval1 gas type
Offval1 gas conc
Offval1 pres mode
Offval1 pres fix val
Offval1 temp mode
Offval1 temp fix val
Offval1 OPL mode
Offval1 OPL fix val
*1:
Description
Selects the type of validation 1 check gas (two-gas measurement only)
Enters the concentration of the validation 1 check gas
Selects the pressure mode for validation 1 execution (*1)
Enters the pressure for when Offval1 pres mode is set to Fixed
Selects the temperature mode for validation 1 execution (*1)
Enters the temperature for when Offval1 temp mode is set to Fixed
Selects the optical path length mode for validation 1 execution (*1)
Enters the process optical path length for when Offval1 OPL mode is set to Fixed
Process parameter: Uses the process parameter value
Fixed value: Set to a fixed value
l Valve
Parameter name
Description
(HART)
Offval1 auto vlv man Selects whether to enable automatic valve control through the SV terminal during
manual validation 1 execution.
Offval1 gas purg time For automatic execution. Set these when performing automatic execution (see section
Offval1 prc purg time 9.8.2).
l Auto time
For automatic execution. Set this when performing automatic execution (see “9.8.2
Configuration”).
9.4.3
Execution
Before starting offline validation, check that the piping and offline validation settings are correct.
Here, offline validation 1 will be used as an example.
Execution menu path:
[HART] “Diagnosis/Service>>Validation>>Manual>>Manual offline val 1”
[YH8000] “
>>Execution>>Validation>>Manual>>Offline Validation 1”
(1) Starting an offline validation
On HART or YH8000, open the above menu, and start offline validation. If automatic
valve control is enabled, a message stating that the valves will be automatically controlled
will appear.*1 In this case, you do not need to manually control the valves during offline
validation.
(2) Purging with check gas
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<9. Inspection and Maintenance>
9-24
An instruction to purge the calibration cell with check gas will appear.*2 If automatic valve
control is disabled, manually control the valves to purge the calibration cell with the check
gas. For safety verification, the standard deviation (stdev) of concentration is displayed
on HART and the concentration trend on the YH8000. Check that the concentration is
stable over a sufficient length of time (5 minutes as a guideline, at least 1 minute) with the
validation cell filled with check gas. Then, execute validation.
(3) Checking the validation result
The validation result is displayed as “PASSED” or “FAILED.” After checking the result,
proceed to the next screen to start purging with the process gas. Or select Retry to return to
(2) and execute validation again.
(4) Discharging the check gas
An instruction to discharge the check gas from the calibration cell will appear.*3 If automatic
valve control is disabled, manually control the valves to purge the calibration cell with
the process gas. For safety verification, the standard deviation (stdev) of concentration
is displayed on HART and the concentration trend on the YH8000. Check that the
concentration is stable, and proceed to the next screen.
(5) Ending validation
The TDLS8000 will exit from validation mode.
*1:
*2:
*3:
[HART]
[YH8000]
[HART]
[YH8000]
[HART]
[YH8000]
Valves are automatically controlled during the validation.
Valve for Check Gas 1 will be opened automatically.
Purge validation cell with check gas.
Purge flow cell with Check Gas 1.
Ensure check gas is completely removed from validation cell and …
Remove Check Gas 1 from flow cell.
NOTE
If the validation fails, the following warning will occur. For the corrective action, see “10.2 Warning
Display and Handling”.
9.4.4
Alarm number
Alarm name
15
Validation Error
Time Chart
The valve operation during manual offline validation execution and the timing when the AO/DO
output switches to Cal/Val mode are shown below. In Cal/Val mode, it is possible to hold the AO
output or specify other settings. For the AO/DO output settings during Cal/Val mode, see “6.4.2
Output Hold” and “6.5.1 DO Contact (DO-1)”.
In the following figure, Valve1 and Valve2 are switched manually by following the instructions on
the operation screen. If automatic valve control is enabled, there is no need for the operator to
switch the valves manually.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9-25
<9. Inspection and Maintenance>
Time
Operator
HART/HMI screen Valve1 Valve2
[Screen operation] Start offline validation 1
OFF
OFF
ON
OFF
AO/DO
mode
Normal output
[Manual valve operation] Purge with check gas
(Wait for the gas concentration to stabilize.)
Check gas
purging
[Screen operation] Touch Execute
(Check validation result.)
[Screen operation] Proceed to the next step (or retry)
Cal/Val
Validation result
[Manual valve operation] Discharge check gas
(Wait for the gas concentration to stabilize.)
Check gas
discharging
OFF
OFF
[Screen operation] Proceed to the next step
Offline validation 1
end
Figure 9.12
Time
Normal output
Valve and AO/DO output for manual offline validation 1
Operator
HART/HMI screen Valve1 Valve2
[Screen operation] Start offline validation 2
OFF
OFF
ON
ON
AO/DO
mode
Normal output
[Manual valve operation] Purge with check gas
(Wait for the gas concentration to stabilize.)
Check gas purging
[Screen operation] Touch Execute
(Check validation result.)
[Screen operation] Proceed to the next step (or retry)
Cal/Val
Validation result
[Manual valve operation] Discharge check gas
(Wait for the gas concentration to stabilize.)
Check gas
discharging
OFF
OFF
[Screen operation] Proceed to the next step
Offline validation 2
end
Figure 9.13
9.5
Normal output
Valve and AO/DO output for manual offline validation 2
Zero Calibration
Zero calibration is a function used to align the zero point in a condition where absolutely none of
the measured components are absorbed by running gas (such as nitrogen) that does not include
the measured components in the region where the laser beam passes through.
Typically zero calibration is performed in an ideal environment before product shipment. In
principle, the TDLS8000 does not have any zero point drift. Therefore, customers normally do not
have to perform zero calibration.
However, if the zero reading is clearly not normal or if you decide that zero calibration is
necessary, perform a zero calibration by paying attention to the following items.
Not meeting the following conditions may adversely affect measurement gas readings.
If you are unclear about how to perform zero calibration, contact your nearest Yokogawa
representative.
Note the following items to perform zero calibration correctly.
• Nitrogen gas concentration meeting the product specifications (99.99%N2 or higher,
depends on the application)
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9-26
<9. Inspection and Maintenance>
Insufficient nitrogen gas concentration may affect the measurement gas concentration
readings.
• The region where the laser beam passes through is adequately filled with nitrogen gas.
If measured gas is mixed, measurement gas concentration readings will be affected.
• There is no optical noise in the region where the laser beam passes through.
Proper zero calibration cannot be performed in a condition where optical noise is present
(for example, if the surface of the process window is clouded). This can affect measurement
gas concentration readings.
• There is no electrical noise in the environment where zero calibration is to be performed.
Proper zero calibration cannot be performed in a condition where electrical noise is present.
This can affect measurement gas concentration readings.
NOTE
• If the purge piping leaks, correct results cannot be obtained.
• Wait at least 1 hour after turning on the power before performing calibration.
9.5.1 Preparation
Follow the instructions in “9.3 Mounting on a Calibration Cell”. The piping method is describe
below.
l Zero+span calibration gas
A piping arrangement that allows span calibration gas to be used in addition to zero calibration
gas. Zero calibration and span calibration can be executed.
Laser unit (LU)
Sensor control unit (SCU)
Flow cell
Purge gas
Purge gas
Process gas
(Stream1)
Valve-2
(SV-2)
Span
(Stream3)
Valve-1
(SV-1)
Zero
(Stream2)
In this arrangement, automatic zero and span calibration can be executed consecutively.
There are two methods to start a consecutive execution. One is by using YH8000, Modbus, or
digital input (semi-automatic execution), and the other is to start at a specified time (automatic
execution). Semi-automatic execution is not possible from HART. For details on automatic and
semi-automatic execution, see “9.8
Automatic and Semi-automatic Execution of Validation
and Calibration”.
For details on consecutive execution, see “9.8.5 Consecutive Automatic Execution”.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9-27
<9. Inspection and Maintenance>
If you want to switch the zero calibration gas stream through automatic valve control using the
TDLS8000 SV terminal, you need to set the TDLS8000 valve usage to Cal/Val. For details on
valve usage, see “6.8.2 Valve Usage Setting”.
9.5.2
Configuration
Zero calibration configuration menu:
[HART] “Detailed setup>>Calibration>>Zero calibration”
[YH8000] “
>>Configuration>Calibration>>Zero Calibration”
The setup parameters required to manually execute zero calibration are indicated for each of the
above submenus (tabs on the YH8000).
l Valve
Parameter name
(HART)
Z-cal auto vlv man
Z-cal gas purg time
Z-cal prc purg time
Description
Selects whether to enable automatic valve control through the SV terminal during
manual zero calibration execution.
For automatic execution. Set these when performing automatic execution (see section
9.8.2).
l Auto time
For automatic execution. Set this when performing automatic execution (see “9.8.2
Configuration”).
9.5.3
Execution
Before starting zero calibration, check that the piping and zero calibration settings are correct.
Execution menu path:
[HART] “Diagnosis/Service>>Calibration>>Manual>>Manual zero cal”
[YH8000] “
>>Execution>>Calibration>>Manual>>Zero Calibration”
(1) Starting zero calibration
On HART or YH8000, open the above menu, and start zero calibration. First, a message
asking you to execute zero calibration carefully will appear.*1 Next, if automatic valve control
is enabled, a message stating that the valves will be automatically controlled will appear.*2
In this case, you do not need to manually control the valves during zero calibration.
(2) Purging with zero calibration gas
An instruction to purge the calibration cell with zero calibration gas will appear.*3 If
automatic valve control is disabled, manually control the valves to purge the calibration
cell with the zero calibration gas. For safety verification, the standard deviation (stdev) of
concentration is displayed on HART and the concentration trend on the YH8000. Check
that the concentration is stable over a sufficient length of time (10 minutes as a guideline,
at least 1 minute) with the validation cell filled with the zero calibration gas. Then, execute
calibration.
(3) Checking the zero calibration result
The result of calibration is displayed as “successful” or “failed.” After checking the result,
proceed to the next screen to start purging with the span calibration gas or process gas. Or
select Retry to return to (2) and execute calibration again.
(4) Purging with process gas
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<9. Inspection and Maintenance>
9-28
An instruction to discharge the zero calibration gas from the calibration cell will appear.*4 If
automatic valve control is disabled, manually control the valves to purge the calibration cell
with the process gas. For safety verification, the standard deviation (stdev) of concentration
is displayed on HART and the concentration trend on the YH8000. Check that the
concentration is stable, and proceed to the next screen.
(5) Ending zero calibration
The TDLS8000 will exit from calibration mode.
*1:
*2:
*3:
*4:
[HART]
[YH8000]
[HART]
[YH8000]
[HART]
[YH8000]
[HART]
[YH8000]
WARN-Please be careful to execute zero calibration.
Are you sure to start manual zero calibration?
Valves are automatically controlled during the calibration.
Valve for Zero Gas will be opened automatically.
Purge calibration cell with zero gas then …
Purge calibration cell with Zero Gas.
Ensure zero gas is completely removed from calibration cell and …
Remove Zero Gas from calibration cell.
NOTE
If the zero calibration fails, the following warning will occur. For the corrective action, see “10.2
Warning Display and Handling”.
Alarm number
Alarm name
16
Zero Cal Error
9.5.4
Time Chart
The valve operation during manual zero calibration execution and the timing when the AO/DO
output switches to Cal/Val mode are shown below. In Cal/Val mode, it is possible to hold the AO
output or specify other settings. For the AO/DO output settings during Cal/Val mode, see “6.4.2
Output Hold” and “6.5.1 DO Contact (DO-1)”.
In the following figure, Valve1 and Valve2 are switched manually by following the instructions on
the operation screen. If automatic valve control is enabled, there is no need for the operator to
switch the valves manually.
Time
Operator
HART/HMI screen Valve1 Valve2
[Screen operation] Start zero calibration
OFF
OFF
ON
OFF
AO/DO
mode
Normal output
[Manual valve operation] Purge with zero calibration gas
(Wait for the gas concentration to stabilize.)
Zero calibration
gas purging
[Screen operation] Touch Execute
(Check the calibration result.)
[Screen operation] Proceed to the next step (or retry)
Cal/Val
Zero calibration result
[Manual valve operation] Discharge zero calibration gas
(Wait for the gas concentration to stabilize.)
Zero calibration
gas discharging
OFF
OFF
[Screen operation] Proceed to the next step
Zero calibration end
Figure 9.14
9.6
Normal output
Valve and AO/DO output for zero calibration
Span Calibration
Zero calibration is a function used to align the concentration calculation result of the TDLS8000
to the concentration of the calibration gas by purging a calibration cell or flow cell with span
calibration gas of known concentration.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9-29
<9. Inspection and Maintenance>
Note the following items to perform span calibration correctly.
• Use gas with accurate concentration for the span calibration gas.
• Perform span calibration with the target region adequately filled with span calibration gas
(purge with calibration gas and check that the reading is adequately stable).
• There is no optical noise in the region where the laser beam passes through.
Proper span calibration cannot be performed in a condition where optical noise is present
(particularly if the surface condition of the process window changes). This can affect
measurement gas concentration readings.
• There is no electrical noise in the environment where span calibration is to be performed.
Proper span calibration cannot be performed in a condition where electrical noise is present.
This can affect measurement gas concentration readings.
NOTE
• If the purge piping leaks, correct results cannot be obtained.
• Wait at least 1 hour after turning on the power before performing calibration.
• Correct measurements may not be obtained if span calibration is performed when the
reading is not stable.
9.6.1 Preparation
Follow the instructions in “9.3 Mounting on a Calibration Cell”. There are two piping methods.
l Zero+span calibration gas
A piping arrangement that allows zero calibration gas to be used in addition to span calibration
gas. Zero calibration and span calibration can be executed.
Laser unit (LU)
Sensor control unit (SCU)
Flow cell
Purge gas
Purge gas
Process gas
(Stream1)
Valve-2
(SV-2)
Span
(Stream3)
Valve-1
(SV-1)
Zero
(Stream2)
In this arrangement, automatic zero and span calibration can be executed consecutively.
There are two methods to start a consecutive execution. One is by using YH8000, Modbus, or
digital input (semi-automatic execution), and the other is to start at a specified time (automatic
execution). Semi-automatic execution is not possible from HART. For details on automatic and
semi-automatic execution, see “9.8
Automatic and Semi-automatic Execution of Validation
and Calibration”.
For details on consecutive execution, see “9.8.5 Consecutive Automatic Execution”.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9-30
<9. Inspection and Maintenance>
l Span calibration gas and two types of process gas
Span calibration gas is connected. The remaining stream is used to switch between two process
gases.
Laser unit (LU)
Sensor control unit (SCU)
Flow cell
Purge gas
Purge gas
Process gas 1
(Stream1)
Valve-2
(SV-2)
Valve-1
(SV-1)
Process gas 2
(Stream2)
Span
(Stream3)
If you want to switch the span calibration gas stream through automatic valve control using the
TDLS8000 SV terminal, you need to set the TDLS8000 valve usage according to the piping
method as shown in the following table. For details on valve usage, see “6.8.2 Valve Usage
Setting”.
Piping
Zero+span calibration gas
Span calibration gas and two types of
process gas
9.6.2
Valve usage setting
Cal/Val
2 Streams & Cal/Val
Configuration
Span calibration configuration menu:
[HART] “Detailed setup>>Calibration>>Span calibration”
[YH8000] “
>>Configuration>Calibration>>Span Calibration”
The setup parameters required to manually execute span calibration are indicated for each of the
above submenus (tabs on the YH8000).
l Parameter
Parameter name
(HART)
S-cal gas type
S-cal gas conc
S-cal pres mode
S-cal pres fix val
S-cal temp mode
S-cal temp fix val
S-cal OPL mode
S-cal OPL fix val
*1:
Description
Selects the type of span calibration gas (two-gas measurement only)
Enters the span calibration gas concentration
Selects the pressure mode for span calibration execution (*1)
Enters the pressure for when S-cal pres mode is set to Fixed
Selects the temperature mode for span calibration execution (*1)
Enters the temperature for when S-cal temp mode is set to Fixed
Selects the optical path length mode for span calibration execution (*1)
Enters the process optical path length for when S-cal OPL mode is set to Fixed
Process parameter: Uses the process parameter value
Fixed value: Set to a fixed value
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<9. Inspection and Maintenance>
9-31
NOTE
For two-gas measurement, you cannot set span calibration simultaneously on two types of gas.
The setting is valid only for the gas type specified by “S-cal gas type,” and span calibration can
be executed on this gas type. To switch the gas to be calibrated, you need to change “S-cal gas
type.” Further, only one type of span calibration gas can be subject to automatic execution.
l Valve
Parameter name
(HART)
S-cal auto vlv man
S-cal gas purg time
S-cal prc purg time
Description
Selects whether to enable automatic valve control through the SV terminal during
manual span calibration execution.
For automatic execution. Set these when performing automatic execution (see section
9.8.2).
l Auto time
For automatic execution. Set these when performing automatic execution (see section 9.8.2).
9.6.3
Execution
Before starting span calibration, check that the piping and span calibration settings are correct.
Execution menu path:
[HART] “Diagnosis/Service>>Calibration>>Manual>>Manual span cal”
[YH8000] “
>>Execution>>Calibration>>Manual>>Span Calibration”
(1) Starting span calibration
On HART or YH8000, open the above menu, and start span calibration. If automatic
valve control is enabled, a message stating that the valves will be automatically controlled
will appear.*1 In this case, you do not need to manually control the valves during span
calibration.
(2) Purging with span calibration gas
An instruction to purge the calibration cell with span calibration gas will appear.*2 If
automatic valve control is disabled, manually control the valves to purge the calibration cell
with the span calibration gas gas. For safety verification, the standard deviation (stdev) of
concentration is displayed on HART and the concentration trend on the YH8000. Check
that the concentration is stable over a sufficient length of time (10 minutes as a guideline,
at least 1 minute) with the validation cell filled with the span calibration gas. Then, execute
calibration.
(3) Checking the span calibration result
The result of calibration is displayed as “successful” or “failed.” After checking the result,
proceed to the next screen to purge with the process gas. Or select Retry to return to (2)
and execute calibration again.
(4) Purging with process gas
An instruction to discharge the span calibration gas from the calibration cell will appear.*3 If
automatic valve control is disabled, manually control the valves to purge the calibration cell
with the process gas. For safety verification, the standard deviation (stdev) of concentration
is displayed on HART and the concentration trend on the YH8000. Check that the
concentration is stable, and proceed to the next screen.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9-32
<9. Inspection and Maintenance>
(5) Ending span calibration
The TDLS8000 will exit from calibration mode.
*1:
*2:
*3:
[HART]
[YH8000]
[HART]
[YH8000]
[HART]
[YH8000]
Valves are automatically controlled during the validation.
Valve for Check Gas 1 will be opened automatically.
Purge calibration cell with span gas then …
Purge calibration cell with Span Gas.
Ensure span gas is completely removed from calibration cell and …
Remove Span Gas from calibration cell.
NOTE
If the span calibration fails, the following warning will occur. For the corrective action, see “10.2
Warning Display and Handling”.
Alarm number
Alarm name
17
Span Cal Error
9.6.4
Time Chart
The valve operation during manual span calibration execution and the timing when the AO/DO
output switches to Cal/Val mode are shown below. In Cal/Val mode, it is possible to hold the AO
output or specify other settings. For the AO/DO output settings during Cal/Val mode, see “6.4.2
Output Hold” and “6.5.1 DO Contact (DO-1)”.
In the following figure, Valve1 and Valve2 are switched manually by following the instructions on
the operation screen. If automatic valve control is enabled, there is no need for the operator to
switch the valves manually.
Time
Operator
HART/HMI screen Valve1 Valve2
[Screen operation] Start span calibration
OFF
OFF
ON
ON
AO/DO
mode
Normal output
[Manual valve operation] Purge with span calibration gas
Span calibration
(Wait for the gas concentration to stabilize.) gas purging
[Screen operation] Touch Execute
(Check the calibration result.)
[Screen operation] Proceed to the next step (or retry)
Cal/Val
Span calibration result
[Manual valve operation] Discharge zero calibration gas
(Wait for the gas concentration to stabilize.)
Span calibration
gas discharging
OFF
OFF
[Screen operation] Proceed to the next step
Span calibration end
Normal output
Figure 9.15
9.7
Calibration Data Record and Restoring
This section explains the function used to view the history of calibration and validation results and
restoring the zero and span calibration data to its original condition.
l Calibration and validation history
You can view up to 10 events using HART and 99 events using the YH8000. For the displayed
history content, see “8.5.6
Cal/Val History Screen”. You can view using the following menu.
[HART] “Diagnosis/Service>>Logbook>>Read cal/val record”
[YH8000] “
>>Log Book>>Cal/Val History”
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<9. Inspection and Maintenance>
9-33
l Restoring calibration data
For zero and span calibration, you can restore past calibration results. You can restore separately
for zero and span. You can select the original data for restoring from the following two types.
• Previous
The calibration data executed previously is restored. When executed, the current calibration
data is saved as past data. Therefore, restoring twice will cause the current calibration data
to return.
• Factory
The factory default calibration data is restored. When executed, the current calibration data
is saved as past data. Therefore, if you restore using “Factory” and then using “Previous,”
the original current data will return.
Execution menu path:
9.8
[HART] “Diagnosis/Service>>Calibration>>Restore”
[YH8000] “
>>Execution>>Calibration>>Restore”
Automatic and Semi-automatic Execution
of Validation and Calibration
There are several methods to perform calibration and validation. One method is manual
calibration and validation, which you execute from the screen. Another method is automatic
calibration and validation, which are executed at a preset time or at preset time intervals. Yet
another method is semi-automatic calibration and validation, which are executed in response to a
start instruction from the YH8000, HART, digital input, or Modbus.
Since valves are controlled automatically in automatic and semi-automatic execution, you
need to set time for preparatory check gas or calibration gas purging. As shown in the following
figure, the time period during which calibration gas (or check gas) purging takes place is called
calibration (validation) time. The subsequent time period during which process gas purging
takes place is called stabilization wait time. The stabilization wait time is the wait time until the
measurements stably return to normal process values. The TDLS8000 is in a Cal/Val state until
the stabilization wait time is completed and holds the AO output. The following figure shows a
remote execution example. Digital input is used to start calibration (validation).
Digital input
Calibration (check)
gas purge
Process purge
AO, DO
Calibration
(Validation) time
Stabilization wait time
AO hold, DO on
NOTE
It is possible to disable the AO hold operation and DO ON operation that take place while
calibration or validation is being executed. For details, see “6.4.2 Output Hold” and “6.5.1 DO
Contact (DO-1)”.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9-34
<9. Inspection and Maintenance>
9.8.1 Preparation
Before automatic execution or semi-automatic execution, connect the piping properly according
to the calibration or validation you want to perform. For the piping diagram, see the calibration
and validation sections.
9.8.2
Configuration
To perform automatic or semi-automatic execution, you need to set certain parameters, which
are shared with manual execution, as well as settings for automatic operation. The settings that
are shared with manual execution are described in the “Configuration” section for validation and
calibration. There are two types of settings for automatic operation. The menu path for accessing
them is the same as that described in the “Configuration” section for validation and calibration.
For online validation, the menu path is as follows.
[HART] “Detailed setup>>Validation>>Online validation #”
[YH8000] “
>>Configuration>>Validation>>Online Validation #”
l Calibration gas or check gas and process gas purge time
The purge time must be set regardless of automatic execution or semi-automatic execution. As
an example, the parameters for online validation 1 are shown below.
Parameter name
(HART)
Onval1 gas purg
time
Onval1 nml purg
time
Parameter name
(YH8000)
Validation gas
Purge time
Normal gas Purge
time
Description
Enters the purge time of online validation 1 check gas. This corresponds
to the validation time.
Enters the normal purge gas purge time for process measurement. This
corresponds to the stabilization wait time.
l Automatic execution settings
To perform automatic execution, you need to set the execution method you want to use. This is
not necessary for semi-automatic execution.
• Parameters for time initiate
As an example, the parameters for online validation 1 are shown below.
Parameter name
(HART)
Onval1 time initiate
Onval1 init date
Onval1 init time
Onval1 day cycle
Onval1 hour cycle
Parameter name
(YH8000)
TIme Initiate
Initial time
Cycle (day)
Cycle (hour)
Description
Enables time initiate of online validation 1
Enters the initial execution date
Enters the initial execution time
Enters the cycle in days for time initiate
Enters the cycle in hours for time initiate
For example, if the initial execution time is 12:00:00 on January 1, 2015, the day cycle is
10, and the hour cycle is 0, the execution will take place at 12:00:00 on January 11, 2015,
12:00:00 on January 21, 2015, and so on.
NOTE
If both the day and hour cycles for time initiate are set to zero, automatic execution takes place
once at the initial execution time.
• If you want to use digital input, see “6.7
Digital Input Settings”.
• If you are using Modbus instructions, you do not need to set the parameters. For the
instruction address, see “11.2
Coil”.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9.8.3
<9. Inspection and Maintenance>
9-35
Execution
Before execution, check that the preparations and settings are correct.
NOTE
If a start request for another calibration or validation overlaps with a calibration or validation
currently in progress, the request will be discarded. For example, if the start time of a time-based
automatic calibration coincides with a manual calibration in progress, the time initiate request is
discarded. Such incidents are recorded in the alarm history.
Since automatic calibration and validation are executed at the specified time cycle, there is no
manual operation to start it. Here, a semi-automatic execution procedure will be explained using
online validation 1 as an example.
Semi-automatic execution menu path:
[HART] “Diagnosis/Service>>Calibration>>Semi-auto”
“Diagnosis/Service>>Validation>>Semi-auto”
[YH8000] “
“
>>Execution>>Calibration>>Semi-Auto”
>>Execution>>Validation>>Semi-Auto”
l HART execution screen
(1) Starting a semi-automatic online validation
Execute “Semi-auto online val 1.” Tap OK.
On the next screen, tap ENTER.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<9. Inspection and Maintenance>
9-36
(2) Purging with check gas
When validation starts, the stream is automatically switched, and the validation cell is
purged with check gas. The purge time is the time specified by “Onval1 gas purg time.”
Tapping ABORT cancels validation.
(3) Discharging the check gas (process purge)
The stream is automatically switched, and the validation cell is purged with normal process
purge gas to discharge the check gas. The purge time is the time specified by “Onval1 nml
purg time.” Tapping ABORT cancels validation.
(4) Checking the validation result
The validation result is displayed as “PASSED” or “FAILED,” and validation ends. If
validation is unsuccessful, a warning will occur (see section 9.2.3). Tap OK to return to the
menu.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<9. Inspection and Maintenance>
9-37
l YH8000 Execution Screen
(1) Starting validation
Execute “Semi-Auto Online Validation 1.” Tap Start to begin.
(2) Purging with the check gas
When validation starts, the stream is automatically switched, and the validation cell is
purged with check gas. The purge time is the time specified by “Validation gas Purge time.”
Tapping Abort cancels validation.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<9. Inspection and Maintenance>
9-38
(3) Discharging the check gas (process purge)
The stream is automatically switched, and the validation cell is purged with normal process
purge gas to discharge the check gas. The purge time is the time specified by “Normal gas
Purge time.” Tapping Abort cancels validation.
(4) Checking the validation result
The validation result is displayed as “PASSED” or “FAILED,” and validation ends. If
validation is unsuccessful, a warning will occur (see section 9.2.3). Tap OK to return to the
configuration menu.
9.8.4
Aborting the Stabilization Wait Time for Automatic or
Semi-automatic Execution
The stabilization wait time during which process purging takes place in an automatic or semiautomatic execution of calibration or validation is a wait time until the measurements are stably
restored. It does not affect the calibration or validation results. Therefore, when you decide that
the process measurements have been stably restored, you can abort the stabilization wait time to
end calibration or validation. This avoids having to wait longer than needed for measurements to
stabilize. For details on the stabilization wait time, see “9.8
Automatic and Semi-automatic
Execution of Validation and Calibration”.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<9. Inspection and Maintenance>
9-39
NOTE
For online validation only, the stabilization wait time cannot be aborted because the
measurements in the normal process condition are used in the validation result. If you abort,
validation itself will be aborted, and the results will not be displayed.
As an example, the procedure to abort process purging for semi-automatic span calibration is
explained below.
(1) Purging with span calibration gas
When you execute semi-automatic span calibration, span calibration gas purging begins.
Note that touching ABORT at this point will abort the span calibration.
(2) Purging with process gas
Immediately after the span calibration gas purge time elapses, the TDLS8000 calculates
the span calibration result. Then, process gas purging takes place. Since span calibration is
already completed at this point, you can view the span calibration result even if you abort the
stabilization wait time. Check that the concentration has stabilized, and then touch ABORT.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9-40
<9. Inspection and Maintenance>
(3) Checking the calibration result
The span calibration result is displayed. Tap OK to return to the menu.
9.8.5
Consecutive Automatic Execution
It is possible to execute zero calibration and span calibration consecutively in automatic
execution or semi-automatic execution. It is also possible to execute offline validation
consecutively using check gas 1 and then check gas 2.
The combinations that can be executed consecutively are shown below.
• Zero calibration and span calibration (zero+span calibration)
• Offline validation 1 and offline validation 2 (offline validation 1+2)
To perform consecutive automatic execution, you need to make the following preparations.
Piping: Connect the piping so that both automatic calibration and validation can
be executed.
Configuration:
Configure the automatic calibration and validation settings.
Automatic execution configuration: Refer to “9.8.2 Configuration”, and configure the
automatic execution settings.
The settings and detailed operation of each consecutive automatic execution are explained
below.
l Zero+span calibration
Piping: Connect the piping for zero+span calibration provided in “9.5.1
Preparation”.
Configuration: Configure the zero calibration (see section 9.5.2) and span calibration
(see section 9.6.2).
Automatic execution configuration:
Configure from the following menu as necessary.
[HART] “Detailed setup>>Calibration>>Zero+Span calibration”
[YH8000] “
>>Configuration>>Calibration>>Zero + Span Calibration”
Semi-automatic execution menu path:
[HART] (Cannot be executed from HART)
[YH8000] “
>>Execution>>Calibration>>Semi-Auto>>Zero + Span Calibration”
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9-41
<9. Inspection and Maintenance>
Time chart: The valve operation during zero+span calibration execution and the timing when
the AO/DO output switches to Cal/Val mode are shown below. The following
figure shows an example for semi-automatic execution.
Time
Operator
YH8000 screen
[Screen operation] Start semi-automatic zero+span calibration
Valve1 Valve2
OFF
OFF
Zero calibration
gas purging
ON
OFF
Span calibration
gas purging
ON
ON
OFF
OFF
AO/DO
mode
Normal output
[Automatic valve operation] Purge with zero calibration gas
(Zero calibration time elapses.)
[Automatic valve operation] Purge with span calibration gas
(Span calibration time elapses.)
Cal/Val
[Automatic valve operation] Discharge zero calibration gas
Span calibration
(Span calibration stabilization wait time elapses.) gas discharging
Span calibration end
Normal output
l Offline validation 1+2
Piping: Connect the piping as described in “Offline validation exclusive” in
section 9.4.1.
Configuration:
Configure the offline validation (see section 9.4.2) for validation 1 and 2.
Automatic execution configuration:
Configure from the following menu as necessary.
[HART] “Detailed setup>>Validation>>Offline validation 1+2”
[YH8000] “
>>Configuration>>Validation>>Offline Validation 1 + 2”
Semi-automatic execution menu path:
[HART] “Diagnosis/Service>>Validation>>Semi-auto>>Semi-auto offline val 1+2”
[YH8000] “
>>Execution>>Validation>>Semi-Auto>>Offline Validation 1 + 2”
Time chart: The valve operation during offline validation 1+2 execution and the timing when
the AO/DO output switches to Cal/Val mode are shown below. The following
figure shows an example for semi-automatic execution.
Time
Operator
HART/YH8000 screen Valve1 Valve2
[Screen operation] Start semi-automatic offline validation 1+2
OFF
OFF
Check gas 1
purging
ON
OFF
Check gas 2
purging
ON
ON
Check gas 2
discharging
OFF
OFF
AO/DO
mode
Normal output
[Automatic valve operation] Check gas 1 purge
(Offline validation 1
validation time elapses.)
[Automatic valve operation] Check gas 2 purge
(Offline validation 2
validation time elapses.)
Cal/Val
[Automatic valve operation] Check gas 2 discharge
(Offline validation 2
stabilization wait time elapses.)
Offline validation 1+2
Result display
Normal output
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9.9
9-42
<9. Inspection and Maintenance>
Analog Input Calibration
This section explains analog-digital conversion calibration of the analog input terminal (AI). Since
the TDLS8000 is calibrated before shipment, you normally do not need to calibrate.
Execution menu path:
[HART] “Diagnosis/Service>>Trim analog channel>>Trim AI-1 (Pres)”
“Diagnosis/Service>>Trim analog channel>>Trim AI-2 (Temp)”
[YH8000] “
“
>>Configuration>>I/O>>Analog Input>>AI-1(Pressure)>>Calibration”
>>Configuration>>I/O>>Analog Input>>AI-2(Temperature)>>Calibration”
The calibration procedure is as follows.
(1) From HART or YH8000, start AI calibration.
(2) Connect a current source to the AI terminal, and apply 4 mA as instructed on the screen.
(3) Check that the analog input is stable, and proceed to the next screen.
(4) Apply 20 mA as instructed on the screen.
(5) Check that the analog input is stable, and proceed to the next screen. Calibration is
complete.
NOTE
If analog input calibration is executed when the analog input is set to process pressure or
temperature, the pressure or temperature during calibration is calculated based on the backup
function. For example, if the pressure’s Backup mode is set to Back value and Backup set value
is set to 101.0 kPa, the pressure at AI-1 during calibration is fixed to 101.0 kPa. For details on the
backup function, see “5.3.3
Setting the Process Pressure”.
9.10
Analog Output Calibration
This section explains digital analog conversion calibration of the analog output terminal (AO).
Since the TDLS8000 is calibrated before shipment, you normally do not need to calibrate.
Execution menu path:
[HART] “Diagnosis/Service>>Trim analog channel>>Trim AO-1 (PV)”
“Diagnosis/Service>>Trim analog channel>>Trim AO-2 (SV)”
[YH8000] “
>>Configuration>>I/O>>Analog Output>>AO-1>>Calibration”
“
>>Configuration>>I/O>>Analog Output>>AO-2>>Calibration”
The calibration procedure is as follows.
(1) Connect ammeter to the AO terminal.
(2) From HART or YH8000, start AO calibration.
(3) A current corresponding to 4 mA will flow. When the measuring instrument reading becomes
stable, enter the measured value.
(4) A current corresponding to 20 mA will flow. When the measuring instrument reading
becomes stable, enter the measured value.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
9-43
<9. Inspection and Maintenance>
(5) Calibration is complete.
NOTE
When analog output calibration is complete, the 20 mA fixed output is released, and normal
analog output returns. At this point, the AO loop check simulation output is also released. For
example, if analog output calibration is executed while AO-1 loop check is in progress, when the
calibration is complete, AO-1 returns to normal output.
9.11
Loop Check
9.12
Alarm History
See “5.4 Loop Check (Simulation output)”.
You can view the history of alarms (faults and warnings) that occurred in the past. In addition, if
an non-alarm event shown in Table 9.1 occurs, it is recorded as a message.
Menu path:
[HART] “Diagnosis/Service>>Logbook>>Read alarm/message record”
[YH8000] “
>>Log Book>>Alarm History”
The information displayed in the alarm history is as follows.
• Times when faults and warnings occur and clear
• Times when messages occur
• Sub numbers of alarm messages (only for certain alarms and messages)
These numbers are used by Yokogawa service representatives for troubleshooting
purposes. The numbers are displayed in HART and the YH8000.
You can view up to 30 events using HART and 99 events using the YH8000. For an explanation
of the YH8000 alarm history screen, see “8.5.5 Alarm History Screen”.
For details on faults and warnings, see “10.
Troubleshooting”.
Items recorded as messages are shown in the following table.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<9. Inspection and Maintenance>
9-44
Table 9.1
No.
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
83
84
85
86
87
88
Message
Power On
Restarted by WDT
Restarted by Power Failure
Laser Module Replaced
Bootloader Updated
Firmware Updated
FPGA Updated
Config File Updated
Backup Config Loaded
Default Config Loaded
Default Firmware Loaded
Default HART config loaded
Reset by External Operation
RTC was Adjusted
Auto Zero Cal was Skipped
Auto Span Cal was Skipped
Auto Validation was Skipped
HMI Connected
HMI Disconnected
HMI Disconnected(recv)
HMI Disconnected(send)
History File was Corrupted
Description
The power was turned on.
Restarted due to a watchdog timeout.
Restarted by a power supply monitoring IC.
Laser module was replaced.
Boot loader was updated.
Firmware was updated.
CIO-FPGA was updated.
The configuration file was updated.
Backup configuration was loaded.
Default configuration was loaded.
Default firmware was loaded.
Default ROM values for HART parameters were loaded.
Restarted by an external instruction.
The real-time clock was synchronized.
Automatic zero calibration start instruction was skipped.
Automatic span calibration start instruction was skipped.
Automatic validation start instruction was skipped.
YH8000 was connected.
YH8000 was disconnected.
YH8000 was disconnected while receiving.
YH8000 was disconnected while sending.
The history file was corrupted.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
10-1
<10. Troubleshooting>
10. Troubleshooting
This chapter explains the faults and warnings that the TDLS8000 may detect. It also explains
how to inspect and restore the TDLS8000 when other problems occurs.
TDLS8000 is composed Laser Unit and Sensor control Unit. And these units are adjusted and
checked in combined. Off-site repair is necessary as a result of troubleshooting, the both units
are sent to factory.
10.1
Fault Display and Handling
A fault occurs when the various types of diagnostic information being monitored by the selfdiagnostics function are clearly abnormal and correct concentration calculation is not possible. It
may signify a malfunction. If a fault occurs, the TDLS8000 output and display will respond in the
following manner.
• The analog output is set to the specified state.
• The fault contact is opened.
• The fault LED (red) lights.
• The alarm indicator blinks on the YH8000 display.
• Alarm information is indicated over HART communication (see section 7.4).
• Alarm information is shown on the SCU display.
The following table shows the fault types and their corrective actions. Alarm numbers are defined
for fault type identification. These numbers are shared among the YH8000, HART, and SCU
displays, even though the abbreviations of the displayed fault names may differ. Depending
on the alarm, a sub number may also be displayed. This number is used by Yokogawa service
representatives for troubleshooting purposes.
NOTE
A fault that occurs is not cleared automatically even when the cause of the fault is eliminated.
Except for a portion of the faults, executing Clear Latched Alarms clears them, but if the cause
of the fault is still present, the fault will occur again. If this happens, the device may have
malfunctioned, so contact your Yokogawa service representative.
Execution menu path:
[HART] “Diagnosis/Service>>>System>>Clear latched alarms”
[YH8000] “
No.
>>Execution>>System>>Clear Latched Alarms”
Displayed name
(HART)
45 Laser Md Temp Low
46
47
48
50
51
Description
Corrective action
The laser module temperature is too Check the laser unit’s ambient temperature.
low.
Contact your Yokogawa service
representative.
Laser Md Temp High The laser module temperature is too Check the laser unit’s ambient temperature.
high.
Contact your Yokogawa service
representative.
Laser Temp Low
The laser temperature is too low.
Contact your Yokogawa service
representative.
Laser Temp High
The laser temperature is too high.
Contact your Yokogawa service
representative.
Peak Center OOR
The absorption peak position is
Contact your Yokogawa service
outside the range.
representative.
Ref Peak Height Low The reference peak height is too low. Contact your Yokogawa service
representative.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
10-2
<10. Troubleshooting>
No.
Displayed name
(HART)
53 Transmission Lost
54 Ref Trans Low
55 R Peak Height High
57 Laser Unit Fail
58 Inter Comm Fail
59 Laser Module Error
60 File Access Error
61 EEPROM Error
62 LU Connect Error
63 FPGA Failure
64 Program Error
*1:
10.2
Description
Corrective action
The transmission is too low to
continue measurements.
Check that the alignment is correct. Check
whether the laser beam is being blocked.
Check whether the process window is
stained. For details, see “9.1
Maintaining the Laser Beam and
Transmission”.
The reference cell transmission is
Contact your Yokogawa service
too low.
representative.
The reference cell peak height is too Contact your Yokogawa service
high.
representative.
The laser unit failed.
Contact your Yokogawa service
representative.
Internal communication error
Contact your Yokogawa service
occurred.
representative.
The laser module failed.
Contact your Yokogawa service
representative. (*1)
File access error.
Contact your Yokogawa service
representative. (*1)
EEPROM error.
Contact your Yokogawa service
representative. (*1)
An error occurred in the laser unit
Check that the laser unit connection cable is
connection.
not loose.
Contact your Yokogawa service
representative.
FPGA failure.
Contact your Yokogawa service
representative. (*1)
Internal error occurred.
Contact your Yokogawa service
representative. (*1)
Cannot be cleared using Clear Latched Alarms. If the fault occurs again even if you restart the TDLS8000, contact your
Yokogawa service representative.
Warning Display and Handling
A warning occurs when the various types of diagnostic information being monitored by the
TDLS8000 self-diagnostics function are outside the normal range. If a warning occurs, the
TDLS8000 output and display will respond in the following manner.
• The analog output is set to the specified state (the factory default hold setting is off).
• DO digital output is generated (for the digital output wiring, see “3.2.5 Wiring Digital
Outputs”).
• The DO LED (yellow) lights.
• The alarm indicator blinks on the YH8000 display.
• Alarm information is indicated over HART communication (see section 7.4).
• Alarm information is shown on the SCU display.
The following table shows the warning types and their corrective actions. Alarm numbers
are defined for warning type identification. These numbers are shared among the YH8000,
HART, and SCU displays, even though the abbreviations of the displayed warning names may
differ. Depending on the alarm, a sub number may also be displayed. This number is used by
Yokogawa service representatives for troubleshooting purposes.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
10-3
<10. Troubleshooting>
No. Displayed name
Description
(HART)
1 Transmission
The transmission is less than the
Low
low limit.
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
19
20
21
Corrective action
Check that the alignment is correct. Check whether
the laser beam is being blocked. Check whether
the process window is stained. Check the low limit
alarm threshold value. For details, see “9.1
Maintaining the Laser Beam and Transmission”.
Pressure Low
The process pressure is less than Check the process gas pressure. Check whether
the low limit.
the gas pressure meter signal is correct. Check
whether the AI range setting is correct. Check the
low limit alarm threshold value. For details, see
“5.3.6 Setting Process Alarms”.
Pressure High
The process pressure is greater
Check the process gas pressure. Check whether
than the high limit.
the gas pressure meter signal is correct. Check
whether the AI range setting is correct. Check the
high limit alarm threshold value. For details, see
“5.3.6 Setting Process Alarms”.
Temperature Low The process temperature is less
Check the process gas temperature. Check
than the low limit.
whether the gas thermometer signal is correct.
Check whether the AI range setting is correct.
Check the low limit alarm threshold value. For
details, see “5.3.6 Setting Process Alarms”.
Temperature
The process temperature is greater Check the process gas temperature. Check
High
than the high limit.
whether the gas thermometer signal is correct.
Check whether the AI range setting is correct.
Check the high limit alarm threshold value. For
details, see “5.3.6 Setting Process Alarms”.
Conc Gas1 Low The component 1 gas
Check the component 1 gas concentration. Check
concentration is less than the low the low limit alarm threshold value. For details, see
limit.
“5.3.6 Setting Process Alarms”.
Check the component 1 gas concentration. Check
Conc Gas1 High The component 1 gas
the high limit alarm threshold value. For details, see
concentration is greater than the
“5.3.6 Setting Process Alarms”.
high limit.
Check the component 2 gas concentration. Check
Conc Gas2 Low The component 2 gas
concentration is less than the low the low limit alarm threshold value. For details, see
“5.3.6 Setting Process Alarms”.
limit.
Check the component 2 gas concentration. Check
Conc Gas2 High The component 2 gas
the high limit alarm threshold value. For details, see
concentration is greater than the
“5.3.6 Setting Process Alarms”.
high limit.
LU Temp Low
The laser unit temperature is too
Check the laser unit’s ambient temperature.
low.
Contact your Yokogawa service representative.
LU Temp High
The laser unit temperature is too
Check the laser unit’s ambient temperature.
high.
Contact your Yokogawa service representative.
SCU Temp Low The sensor control unit
Check the sensor control unit’s ambient
temperature is too low.
temperature. Contact your Yokogawa service
representative.
SCU Temp High The sensor control unit
Check the sensor control unit’s ambient
temperature is too high.
temperature. Contact your Yokogawa service
representative.
Validation
Validation is required for verifying Execute validation. Or, if you confirm that validation
Required
the measurement accuracy.
is not required, clear the alarm. (*1)
Validation Error Validation failed.
Verify the check gas. Check whether the validation
settings are correct. For details, see “9.2
Online Validation” and “9.4 Offline Validation”. (*1)
Zero Cal Error
Zero calibration failed.
Check the zero calibration gas. Check whether the
zero calibration settings are correct. For details, see
“9.5
Zero Calibration”. (*2)
Span Cal Error Span calibration failed.
Check the span calibration gas. Check whether the
span calibration settings are correct. For details,
see “9.6 Span Calibration”. (*2)
Non Process
The non-process reference peak Check whether the purge gas is running. Check the
Alarm
height is too low.
purge gas concentration. Contact your Yokogawa
service representative.
AI-1 (Pres) Low The AI-1 (pressure) input current is Check the process gas pressure. Check whether
less than 4 mA.
the gas pressure meter signal is correct. Check
whether the AI range setting is correct. For details,
see “6.3 Analog Input Settings”.
AI-1 (Pres) High The AI-1 (pressure) input current is Check the process gas pressure. Check whether
more than 20 mA.
the gas pressure meter signal is correct. Check
whether the AI range setting is correct. For details,
see “6.3 Analog Input Settings”.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
10-4
<10. Troubleshooting>
No. Displayed name
Description
(HART)
22 AI-2 (Temp) Low The AI-2 (temperature) input
current is less than 4 mA.
23 AI-2 (Temp) High The AI-2 (temperature) input
current is more than 20 mA.
24 External Alarm
An alarm triggered by digital input
occurred.
25 Clock Adj
The real-time clock is not
Required
synchronized.
26 Setting File Error Restored from backup due to a
setup file corruption.
27 Calib File Error Restored from backup due to a
calibration file corruption.
49 Detect Signal
The detector signal level is too
High
high.
52 Absorption High The absorption signal level is too
high.
56 Outlier Reject
The detector signal level is
Lmt
abnormal.
Corrective action
Check the process gas temperature. Check
whether the gas thermometer signal is correct.
Check whether the AI range setting is correct. For
details, see “6.3 Analog Input Settings”.
Check the process gas temperature. Check
whether the gas thermometer signal is correct.
Check whether the AI range setting is correct. For
details, see “6.3 Analog Input Settings”.
Check the external alarm status.
Set the current time.
Configure the settings again, and restart.
Calibrate again, and restart.
Contact your Yokogawa service representative.
Contact your Yokogawa service representative.
Contact your Yokogawa service representative.
*1:
For validation alarms, you can manually clear the alarm without re-executing validation.
Menu path:
[HART] “Diagnosis/Service>>Validation>>Clear val alarms”
[YH8000] “
>>Execution>>Validation>>Clear Validation Alarm”
*2:
For calibration alarms, you can manually clear the alarm without re-executing calibration.
Menu path:
[HART] “Diagnosis/Service>>Calibration>>Clear cal alarms”
[YH8000]
10.3
“
>>Execution>>Calibration>>Clear Calibration Alarm”
Handling Degraded Laser Transmission
For the TDLS8000 to operate normally, the optimal level of laser beam needs to reach the sensor
control unit (SCU).
The following phenomena can cause the laser beam level to degrade. These factors may occur
separately or together.
●
Optical axis error: Degradation of received light level due to optical axis misalignment
• The laser beam is not directed at the detector window.
• The SCU has not be aligned with the laser incident beam.
• Misaligned flange or nozzle is blocking the laser beam.
●
Clogging: The opening where the laser beam travels through is blocked or is unclean.
• Dust has accumulated inside nozzles or pipes, blocking the laser beam.
• Stain or foreign substances adhering to the process window are attenuating the laser beam
level.
●
Particles: Dust in the process gas is attenuating the laser beam level.
• Smoke concentration, opacity, or particle concentration is extremely high, and not enough
laser beam is reaching the SCU.
●
Laser degradation: The output power of the laser element itself has degraded.
• The laser light source has degraded or malfunctioned, and not enough laser beam is being
emitted.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
10-5
<10. Troubleshooting>
n Improving transmission
This section provides corrective actions for when the transmission is lost or reduced after
installing the TDLS8000.
(1) Checking for mechanical blockage
If process isolation valves or the like are used to shut off the process gas for periodic
maintenance, check that the valve for the LU or SCU is not closed. Since the laser beam will be
blocked if the valve is closed, open it. If the transmission does not improve even when the valve
is opened, other factors may be causing the problem.
(2) Adjusting the optical axis
If the laser beam is not shut off but the transmission is low, double check that the optical axis is
adjusted correctly. As described in “9.1 Maintaining the Laser Beam and Transmission”,
normally optical axis adjustment and transmission calibration are performed after the TDLS8000
is installed. But if the process gas temperature is high, the optical axis may diverge from the initial
adjustment due to a deformation in the duct or the like causing the process flange or nozzle to be
misaligned.
If optical axis readjustment is necessary, do so by referring to “3.3 Optical Axis Adjustment”.
Laser unit (LU)
Sensor control unit (SCU)
Measured gas
(a) When optical axis adjustment is complete (optical axis is correct)
Laser unit (LU)
Sensor control unit (SCU)
Measured gas
Measured gas
(b) When optical axis adjustment is necessary (optical axis is not correct)
Figure 10.1
Conditions in which optical axis adjustment is necessary
If the transmission does not improve even when the optical axis adjustment described in section
3.3 is executed, other factors may be causing the problem.
(3) Checking and correcting the opening to the process
If no improvement is seen even when the measures described in (1) and (2) above are taken,
the condition of the opening to the process may be causing the problem. For example, if dust or
the like in the process gas accumulates inside a process flange or insertion tube, it may block the
laser beam.
Remove the TDLS8000 from the alignment flange, and check that deposits have not
accumulated inside the process flanges and insertion tubes. If deposits are found, remove it
according to the following procedure.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
10-6
<10. Troubleshooting>
Figure 10.2
Example of deposits in the opening
(a) Turn the power off.
(b) Stop the LU and SCU analyzer purging.
(c) Remove the LU and SCU from the alignment flange.
(d) Check the condition of the process opening through the process window.
If accumulation is found, take the following corrective action.
(e) Remove the process interface (alignment flange or process isolation flange).
(f)
If insertion tubes are used, remove them from the process. Then, remove the deposits. If
deposits are found in the process flanges, remove them.
(g) Attach the process interface to the process again.
(h) Install the TDLS8000, supply purge gas and power, and turn on the power.
If you removed the alignment flange, readjust the optical axis according to “3.3 Optical Axis
Adjustment”.
If the transmission does not improve even when deposits are removed, other factors may be
causing the problem.
(4) Checking and correcting laser output degradation or loss
If the transmission is low or at zero even when the corrective measures described in (1), (2), and
(3) are taken, the laser output may be degraded or the laser itself may be malfunctioning.
To check the laser output power, follow the procedure below.
(a) Turn off the TDLS8000.
(b) Remove the LU and SCU from the process.
CAUTION
Be sure to turn off the TDLS8000 before removing it from the process.
Removing the TDLS8000 without turning off the power constitutes a risk of laser beam entering
your eye.
(c) Connect the LU and SCU directly as shown in Figure 10.3.
Connect it so that the laser beam is not blocked.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
10-7
<10. Troubleshooting>
Laser Unit
Figure 10.3
Sensor Control Unit
Direct connection of the LU and SCU
(d) Turn the power on, and check the transmission.
If the transmission improved as a result of connecting the LU and SCU directly, slightly
adjust the angle to check whether the signal intensifies. If the transmission is zero even if
you adjust the angle between the LU and SCU, the device is broken. Please consult with
Yokogawa.
10.4
Process Window Replacement
10.4.1
Replacement Parts (Process window)
If the stain on the surface of a process window does not come off even if you clean it according to
section 9.1.2 or if the surface has corroded due to corrosive gas such as hydrogen fluoride, you
need to replace the process window. Replace it according to the procedures in sections 10.3.2
and 10.3.3. When you replace a process window, be sure to also replace the O-ring.
If you need to replace a process window, prepare the relevant parts in Table 10.1.
Table 10.1
No.
1
2
3
4
5
6
7
8
9
10
11
12
Parts no.
K9772RA
K9772RB
K9772RC
K9772RD
K9772RE
K9772TH
K9775EA
K9775EB
K9775EC
K9775ED
K9775EE
K9775GE
Replacement parts
Parts name
Process window unit
Process window unit
Process window unit
Process window unit
Process window unit
O-ring
Sapphire window unit
Sapphire window unit
Sapphire window unit
Sapphire window unit
Sapphire window unit
O-ring
Purpose *
For alignment flanges (for -X1, -X2)
For alignment flanges (for -H1)
For alignment flanges (for -A1)
For alignment flanges (for -H3)
For alignment flanges (for -C3, -C4)
For process window attached to alignment flange
For process isolation flanges or flow cells (for -X1, -X2)
For process isolation flanges or flow cells (for -H1)
For process isolation flanges or flow cells (for -A1)
For process isolation flanges or flow cells (for -H3)
For process isolation flanges or flow cells (for -C3, -C4)
For process window attached to process isolation flange
or attached to flow cell
For process window attached to flow cell
Quantity
1
or
2
*:
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<10. Troubleshooting>
10.4.2
10-8
Process Window Replacement Procedure (Alignment
flange)
The procedure for replacing the process window mounted on an alignment flange is provided
below.
NOTE
Before removing the process window, check that the process is completely stopped or that the
TDLS8000 is isolated from the process and no process gas will be discharged. Never remove a
process window while the process is running as it is extremely dangerous.
CAUTION
Be careful in handling the process window as it is made of optical glass.
(1) Turn the power off.
(2) Stop the purge gas.
(3) Remove the purge piping.
(4) Remove the TDLS8000 from the process.
(If necessary, separate it completely from the process such as by using a process isolation
valve.)
(5) Check the stained area of the process window, and remove the relevant process window.
(6) Loosen the four M4 screws with hexagonal holes on the process window holder installed in
the alignment flange, and remove the process window.
(7) Install a new process window. Replace the O-ring also.
The position of the O-ring is indicated in Figure 10.4. Firmly mount the O-ring in the O-ring
groove.
Alignment flange
Process window
O-ring
Hole with a hexagonal hole
Figure 10.4
Replacing the process window and O-ring of an alignment flange
(8) Pay attention to the orientation of the process window. Install it in the same orientation as
before.
Tighten the screws evenly.
(9) After installing the process window, install the TDLS8000 for use.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
10.4.3
10-9
<10. Troubleshooting>
Process Window Replacement Procedure (Process
isolation flange)
The procedure for replacing the process window mounted on a process isolation flange is
provided below.
NOTE
Before removing the process window, check that the process is completely stopped or that the
TDLS8000 is isolated from the process and no process gas will be discharged. Never remove a
process window while the process is running as it is extremely dangerous.
CAUTION
Be careful in handling the process window as it is made of optical glass.
(1) Turn the power off.
(2) Stop the purge gas.
(3) Remove the purge piping.
(4) Remove the TDLS8000 from the process.
(If necessary, separate it completely from the process such as by using a process isolation
valve.)
(5) Remove the alignment flange.
(6) Check the stained area of the process window of the process isolation flange, and remove
the process window.
(7) Loosen the six M5 screws with hexagonal holes on the process window holder installed in
the process isolation flange, and remove the process window.
Make sure not to drop the process window. There is no mechanism from preventing the
screws with hexagonal holes or the washers from falling. Make sure not to drop or lose
them.
(8) Install a new process window. Replace the O-ring also.
The position of the O-ring is indicated in Figure 10.5. Firmly mount the O-ring in the O-ring
groove.
Process isolation flange
Process window
O-ring
Washer
Hole with a hexagonal hole
Figure 10.5
Replacing the process window and O-ring of a process isolation flange
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
10-10
<10. Troubleshooting>
(9) Pay attention to the orientation of the process window. Install it in the same orientation as
before.
Tighten the screws evenly.
(10) After mounting the process window, install the alignment flange and then the TDLS8000.
10.4.4
Since you removed the alignment flange from the process, readjust the optical axis.
Process Window Replacement Procedure (Flow cell)
The procedure for replacing the process window mounted on a flow cell is provided below.
There are two types of process windows used in a flow cell: a process window mounted on
the alignment flange for the flow cell and a process window mounted on the flow cell itself. The
process window mounted on the flow cell itself can only be replaced by removing the alignment
flange for the flow cell.
NOTE
Before removing the process window, check that the process is completely stopped or that the
TDLS8000 is isolated from the process and no process gas will be discharged. Never remove a
process window while the process is running as it is extremely dangerous.
CAUTION
Be careful in handling the process window as it is made of optical glass.
(1) Turn the power off.
(2) Stop the purge gas.
(3) Remove the purge piping.
(4) Remove the TDLS8000 from the flow cell.
(Be sure to separate the flow cell completely from the process in advance such as by using
a process isolation valve.)
(5) The procedure to replace the process window mounted on the alignment flange is similar to
that explained in 10.4.2.
(6) To remove the process window mounted on the flow cell, remove the alignment flange
mounted on the flow cell.
(7) Check the stained area of the process window of the flow cell, and remove the process
window.
(8) Loosen the six M5 screws with hexagonal holes on the process window holder installed in
the flow cell, and remove the process window.
Make sure not to drop the process window. There is no mechanism from preventing the
screws with hexagonal holes or the washers from falling. Make sure not to drop or lose
them.
For details on how to install and remove the process window, see Figure 10.6.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
10-11
<10. Troubleshooting>
Hole with a hexagonal hole
Process window
Flow cell
Washer
Figure 10.6
O-ring
for flow cell
Removing and installing the process window of a flow cell
(9) Install a new process window. Replace the O-ring for flow cell also.
The position of the O-ring is indicated in Figure 10.7.
Flow cell
Hole with a
hexagonal hole
(M5x6)
Washer
Flow cell alignment
flange
Hole with a
hexagonal hole
(M4x4)
O-ring
O-ring
Hole with a hexagonal hole
(M10x4)
Process window
Figure 10.7
O-ring for flow cell
Washer
Replacing the process window and O-ring of a flow cell
(10) Pay attention to the orientation of the process window. Install it in the same orientation as
before.
Tighten the screws evenly.
(11) After mounting the process window, install the alignment flange for the flow cell and then
the TDLS8000. Since you removed the alignment flange for the flow cell from the process,
readjust the optical axis.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
10.5
10-12
<10. Troubleshooting>
Fuse Replacement
(1) To safely replace the fuse, shut off the external circuit breaker to stop the power supply to
the TDLS8000.
(2) Remove the fuse from the fuse holder. Using a flat-blade screwdriver that matches the
holder cap, turn the cap 90 degrees counterclockwise.
Then you will be able to remove the fuse with the cap.
(3) Check that the rating of the new fuse is correct, place it in the fuse cap, and insert the cap in
the holder. Using a flat-blade screwdriver, turn the cap 90 degrees clockwise while pressing
down.
Be careful because the fuse for the SCU is 3.15 A and that for the LU and YH8000 is 2.5 A.
(4) If the new fuse blows immediately, there may be a problem with the circuitry. Contact your
Yokogawa representative.
Fuse (2.5 A)
Laser unit (LU)
Fuse (3.15 A)
Sensor control unit (SCU)
Fuse (2.5 A)
YH8000 HMI unit
Figure 10.8
Fuse replacement
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
10.6
<10. Troubleshooting>
10-13
Communication Interruption during Manual
Calibration and Validation
If the communication between HART or YH8000 and the TDLS8000 is disconnected while
performing manual calibration or validation from HART or YH8000, take the following corrective
action.
l HART
See “7.5.3
Aborting Calibration and Validation”.
l YH8000
(1) Reconnect.
to enter the TDLS8000 configuration screen. The screen for the calibration or
(2) Tap
validation in progress automatically recovers. You can continue the calibration or validation.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
Blank Page
<11. Modbus>
11-1
11. Modbus
Modbus protocol can be used for TDLS8000 DCS communication. This section explains the
Modbus communication specifications that apply to the TDLS8000.
The main uses of Modbus communication on the TDLS8000 are shown below. Only a portion of
the TDLS8000 configuration function is supported.
• Checking measured values, I/O, and alarms
• Executing calibration, validation, and clock setting
• A portion of configuration functions (setting the current stream, inputting the temperature
and pressure)
11.1
Communication Specifications
The TDLS8000 can be used as a Modbus slave device. Modbus communication is possible by
connecting to a master device via Ethernet cable.
Communication standard
Number of sessions (max.)
Protocol
Port number
11.1.1
Ethernet
2
Modbus/TCP
502
Message Structure
The communication message structure is shown below. The first seven bytes are the Modbus/
TCP header.
Bytes
1
2
3
4
5
6
7
8
9
:
Transaction ID
Protocol ID
Modbus/TCP header
Message length
Unit ID
Function
Data
l Transaction ID
Data assigned by the master device to manage transactions. Slave devices simply return the
received value.
l Protocol ID
Fixed at zero.
l Message length
Data byte length after the unit ID.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
11-2
<11. Modbus>
l Unit ID
Don’t care for Modbus/TCP. Slave devices simply return the received value.
l Function
The supported function numbers are listed in the following table.
Function no.
Function
Type
Max. number of data points per transaction
1
Read coils
Bit
2000 points
2
Read the input relays
Bit
2000 points
3
Read hold registers
Word
125 points
4
Read the input registers
Word
125 points
5
Write to a single coil
Bit
1 point
6
Write to a single hold register
Word
1 point
16
Write to hold registers
Word
123 points
43
Read device information
ASCII string
(*1)
*1:
The following parameters, which are basic device ID parameters (in the basic category) are read by function 43.
ID
Object name
0x00 VenderName
0x01 ProductCode
Meaning
Vendor name
Product code
0x02 MajorMinorRevision Revision number
Value
"YOKOGAWA"
"TDLS8000"
"[Device Revision]-[Software Revision]"
Example: "01-1.01.01"
l Data
There are two types of data: “coil/relay” in unit of bits and “register” in unit of 16 bits. Data
attributes and data addresses are shown in the following table.
Type
Bit
Register
*1:
11.1.2
Attribute
W
R
R
W
Modbus name
Coil
Input relay
Input register
Hold register
Address (*1)
0XXXX
1XXXX
3XXXX
4XXXX
Application
Instruction
Status
Measured value
Valve control, temperature/pressure input
XXXX: 0001 to 9999
Slave Response
Function and subsequent content of response messages vary depending on whether there are
errors in instruction messages.
l Normal response
In the case of writing to a single coil or single hold register, the slave device returns the same
message as the instruction message. In the case of a read function, the read data is added to the
function in the response message. If an address in which no data is assigned is read, zero, not
an error, is returned as the read data.
l Error response
If there is an error in the instruction message, the slave device returns an error response without
executing the instruction. In an error response, the slave device returns the value obtained by
adding 128 to the instruction function as the error function. Therefore, the master device can
check the function in the response message to determine whether an instruction has been
accepted normally. If the master device determines an error has occurred, it can find out the
details by checking the error code.
The message structure from the function and beyond in an error response is as follows.
Error function (instruction function + 128)
Error code
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
11-3
<11. Modbus>
The error code details are provided below.
Error code
Description
01
02
03
Function code error (nonexistent function)
Coil, input relay, or register address error (out of range)
Coil, input relay, or register data number error (out of range)
During instruction message execution, an error which is the slave device
cannot execute occurs.
Example: Writing not possible because maintenance is in progress
Command error
Example: Write-data is out of range.
06
07
11.2 Coil
Coil name
Automatic zero calibration execution
Automatic span calibration execution
Automatic offline validation 1 execution
Automatic offline validation 2 execution
Automatic online validation 1 execution
Automatic online validation 2 execution
Time set instruction
Automatic zero+span calibration
execution
Automatic offline validation 1+2
execution
Address
Action performed when “1” is set
00001
00002
00004
00005
00006
00007
00008
Remotely execute an automatic zero calibration
Remotely execute an automatic span calibration
Remotely execute automatic offline validation 1
Remotely execute automatic offline validation 2
Remotely execute automatic online validation 1
Remotely execute automatic online validation 2
Set the hold register time value (40201 to 40206)
00009
Remotely execute an automatic zero+span calibration
00010
Remotely execute automatic offline validation 1+2
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
11.3
<11. Modbus>
11-4
Input relay
Input relay name
Analyzer error
Maintenance in progress
AO-1, 2 fixed output
AO-1 fixed output
AO-2 fixed output
Zero calibration in progress
Span calibration in progress
Offline validation in progress
Online validation in progress
Warming up
Normal measurement in progress
Measurement update notification
Alarm update
Instruction failure update
Manual Zero Calibration
Manual Span Calibration
Automatic Zero Calibration
Automatic Span Calibration
Manual Offline Validation 1
Manual Offline Validation 2
Manual Online Validation 1
Manual Online Validation 2
Automatic Offline Validation 1
Automatic Offline Validation 2
Automatic Online Validation 1
Automatic Online Validation 2
Digital input state
Digital output state
Address
Description
10001 Alarm occurring when set to 1 (refer to address
10101 and beyond for the alarm details)
10002 Maintenance in progress when set to 1
10003 Both AO-1 and 2 fixed output when set to 1
10004 AO-1 fixed output in progress when set to 1
10005 AO-2 fixed output in progress when set to 1
10006 Zero calibration in progress when set to 1
10007 Span calibration in progress when set to 1
10009 Offline validation in progress when set to 1
10010 Online validation in progress when set to 1
10013 Warming up when set to 1
10014 Normal measurement in progress when set to 1
10015 Set to 1 after measurement is updated. Reading
this address resets the value to 0. (*1)
10016 Set to 1 when a new alarm occurs or when an
alarm is cleared. Reading this address resets the
value to 0. (*1)
10017 Set to 1 when an instruction by a coil fails.
Reading this address or a successful next
instruction resets the value to 0. (*1)
10031 Calibration (validation) in progress when set to 1
10032
10034
10035
10037
10038
10039
10040
10041
10042
10043
10044
10051 DI-1 contact state (0: Open, 1: Closed)
10052 DI-2 contact state (0: Open, 1: Closed)
10061 DO contact (DO-1) state (0: Off, 1: On)
10062 Fault contact (DO-2) state (0: Off, 1: On)
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
<11. Modbus>
Input relay name
Warning: Transmission low (AL-1)
Warning: Process pressure low (AL-2)
Warning: Process pressure high (AL-3)
Warning: Process temperature low (AL-4)
Warning: Process temperature high (AL-5)
Warning: Concentration gas1 low (AL-6)
Warning: Concentration gas1 high (AL-7)
Warning: Concentration gas2 low (AL-8)
Warning: Concentration gas2 high (AL-9)
Warning: Laser unit temperature low (AL-10)
Warning: Laser unit temperature high (AL-11)
Warning: Sensor control unit temperature low
(AL-12)
Warning: Sensor control unit temperature high
(AL-13)
Warning: Validation required (AL-14)
Warning: Validation failure(AL-15)
Warning: Zero calibration error (AL-16)
Warning: Span calibration error (AL-17)
Warning: Non process alarm (AL-19)
Warning: AI Ch1 (pressure) low (AL-20)
Warning: AI Ch1 (pressure) high (AL-21)
Warning: AI Ch2 (temperature) low (AL-22)
Warning: AI Ch2 (temperature) high (AL-23)
Warning: External alarm (AL-24)
Warning: Clock adjustment required (AL-25)
Warning: Setting file corrupted (AL-26)
Warning: Calibration file corrupted (AL-27)
Fault: Laser module temperature low (AL-45)
Fault: Laser module temperature high (AL-46)
Fault: Laser temperature low (AL-47)
Fault: Laser temperature high (AL-48)
Warning: Detector signal high (AL-49)
Fault: Peak center out of range (AL-50)
Fault: Reference peak height low (AL-51)
Warning: Absorption too high (AL-52)
Fault: Transmission lost (AL-53)
Fault: Reference transmission low (AL-54)
Fault: Reference peak height high (AL-55)
Warning: Outlier rejection limit (AL-56)
Fault: Laser unit failure (AL-57)
Fault: Internal communication failure (AL-58)
Fault: Laser module error (AL-59)
Fault: File access error (AL-60)
Fault: EEPROM error (AL-61)
Fault: Laser Unit Connection Error (AL-62)
Fault: FPGA Failure (AL-63)
Fault: Program error (AL-64)
*1:
11-5
Address
Description
10101 Alarm occurring when set to 1
10102 * (AL-##) in the name column denotes the alarm
number.
10103
10104
10105
10106
10107
10108
10109
10110
10111
10112
10113
10114
10115
10116
10117
10119
10120
10121
10122
10123
10124
10125
10126
10127
10145
10146
10147
10148
10149
10150
10151
10152
10153
10154
10155
10156
10157
10158
10159
10160
10161
10162
10163
10164
If this address is read from two sessions, the first access has priority.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
11.4
Hold register
Name
Valve stream setting
Temperature input
value
Pressure input value
Time setting (year)
Time setting (month)
Time setting (day)
Time setting (hour)
Time setting (minute)
Time setting (second)
*1:
*2:
*3:
11-6
<11. Modbus>
Address
40001
Setting details
Sets the current stream
0: Stream 1
1: Stream 2
2: Stream 3
*: Note that via Modbus, exclusion check on the writing of the
current stream according to the valve usage setting is not
performed. For details on exclusion, see “6.8.2 Valve Usage
Setting”.
*: Writing is not possible when maintenance is in progress.
40101, 40102 Temperature input value via Modbus, IEEE754 float format (*1)
The unit follows to the temperature unit setting.
*: Writing is possible even when maintenance is in progress.
40103, 40104 Pressure input value via Modbus, IEEE754 float format (*1)
The unit follows to the pressure unit setting.
*: Writing is possible even when maintenance is in progress.
40201
RTC setting date/time (year) based on 2000 (2015 is expressed
as 15) (*2) (*3)
40202
RTC setting date/time (month) 1 to 12 (*2) (*3)
40203
RTC setting date/time (day) 1 to 31 (*2) (*3)
40204
RTC setting time (hour) 0 to 23 (*2)
40205
RTC setting time (minute) 0 to 59 (*2)
40206
RTC setting time (second) 0 to 59 (*2)
IEEE754 float format (in 2 registers, In the order upper 16 bits and then lower 16 bits)
Write the both upper and lower bits together.
Apply the settings using the coil “time setting instruction”.
Write the year, month, and day in order from the highest address.
NOTE
When inputting the temperature value or pressure value via Modbus, set the input unit the same
as the TDLS8000 unit. If input using a different unit, the concentration reading will not be output
correctly.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
11.5
<11. Modbus>
Input register
Input register name
Concentration value
Transmission value
Temperature value
Pressure value
AI value
AO value
Calibration/validation
execution state
Active alarm state value
Current time (year)
Current time (month)
Current time (day)
Current time (hour)
Current time (minute)
Current time (second)
*1:
11-7
Address
Description
30001, 30002 Component 1 gas concentration value, IEEE754 float format (*1)
The unit follows to the component 1 gas setting.
30003, 30004 Component 2 gas concentration value, IEEE754 float format (*1)
The unit follows to the component 2 gas setting.
30007, 30008 Transmission [%], IEEE754 float format (*1)
30011, 30012 Temperature value, IEEE754 float format (*1)
The unit follows to the temperature unit setting.
30015, 30016 Pressure value, IEEE754 float format (*1)
The unit follows to the pressure unit setting.
30019, 30020 AI-1 current value [mA], IEEE754 float format (*1)
30021, 30022 AI-2 current value [mA], IEEE754 float format (*1)
30025, 30026 AO-1 current value [mA], IEEE754 float format (*1)
30027, 30028 AO-2 current value [mA], IEEE754 float format (*1)
30031
A value indicating the calibration/validation execution state
Value
Calibration state
0 Not in progress
1 Zero calibration
2 Span calibration
5 Offline validation
6 Online validation
30035 to 30038 Indicates active alarm states.
A value in unsigned long format (Big-endian arrangement in four
registers). Bit numbers corresponding to alarm numbers in which
warning or fault is occurring are set to 1. When multiple alarms are
occurring, they are expressed as a sum of the bits.
Example:
The read value when transmission low (alarm
number 1) and transmission lost (alarm number 53) are
occurring is 0x10000000000001:
30035: 0x0010$3
30036: 0x0000$3
30037: 0x0000$3
30038: 0x0001$3
30201
RTC Current time (year) based on 2000
30202
RTC Current time (month) 1 to 12
30203
RTC Current time (day) 1 to 31
30204
RTC Current time (hour) 0 to 23
30205
RTC Current time (minute) 0 to 59
30206
RTC Current time (second) 0 to 59
IEEE754 float format (in 2 registers, In the order upper 16 bits and then lower 16 bits)
Read the both upper and lower bits together.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
Blank Page
App.1-1
Appendix 1 Constructing Unit
Connection Cables
Use the cables of other selling to connect between the sensor control unit (SCU) and the laser
unit (LU). This section explains how to terminate this cable.
NOTE
To maintain the TDLS8000 performance and functionality, be sure to use the cables of other
selling.
l Required components and tools
• Wire cutter
• Wire stripper
• Round crimp-on terminals (for M4 screws, nominal cross sectional area of wire 5.5 mm2), 2 pcs.
•
•
•
FV5.5-S4 by J.S.T. Mfg. Co.,Ltd. or 324915 by TE Connectivity, or equivalent
Crimp tool for round crimp-on terminals
Heat shrink tubes
Inner diameter 6 mm, length approx. 110 mm, 2 pcs.
Inner diameter 16 mm, length approx. 50 mm, 2 pcs.
Heating gun (for shrinking heat shrink tubes)
l Cable specifications
• Manufacturer and model Belden 1475A
• Number of pairs 4 pairs (cover colors, black and white)
Pair numbers are indicated on the cover of each wire (ONE, TWO,
THREE, FOUR)
• Shield
Shield for each pair and shield for the whole cable
• Wire diameter
AWG18
• Cable outer diameter
Approx. ø12 mm
• Flame resistance
FT4
• Operating temperature range
-30°C to +105°C
• Usage environment
Indoor/outdoor
For detailed cable specifications, visit the Belden website.
Select cables with the appropriate length from “2.2.5 Unit Connection Cable” according to the
TDLS8000 installation conditions and adjust the length if necessary.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.1-2
l How to terminate the cables
Terminate both ends of the cables in the same manner.
(1) Remove about 120 mm of the outer cover from the ends of the cables.
Be careful not to damage the wires, shields, and the like in the cables.
Outer covering
Metallic foil
About 120 mm
Removing the outer cover
(2) Remove the metallic shield foil from the outside of the cable.
Each twisted-pair is
further wrapped in metallic foil.
Outer covering
Orange wire
Universal ground wire
Removing the external metallic shield foil
(3) Remove the metallic shield foil from each twisted-pair (four pairs) and the orange wire.
Outer covering
Ground wire of each twisted-pair
Universal ground wire
Removing the metallic shield foil from each twisted-pair and the orange wire
(4) Bundle the ground wire of each twisted-pair (four pairs) and the universal ground wire
together.
Each twisted-pair consists of a black wire and a white wire. On the cover of each wire,
the pair number (ONE, TWO, THREE, FOUR) is marked. If necessary, apply additional
markings for twisted-pair identification.
Outer covering
Bundle the ground wire of
each twisted-pair and the
universal ground wire together.
Bundling ground wires
(5) Cover the bundled ground wires with a heat shrink tube (6 mm inner diameter, about 110
mm in length), and apply heat treatment. Keep about 10 mm of the ends of the ground wires
exposed from the heat shrink tube.
Then, as shown in the figure, cover the entire cable with a heat shrink tube (16 mm inner
diameter, about 50 mm in length), and apply heat treatment.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.1-3
Outer covering
Cover with a heat shrink tube.
Heat shrink tube
About 110 mm
Expose about 10 mm of the end of
the ground wires.
Cover with a heat shrink tube.
Heat shrink tube
About 50 mm
Overlap from the outside cover
About 20 mm
Covering with heat shrink tubes
(6) Attach a round crimp-on terminal (for M4 screws, nominal cross sectional area of wire 5.5
mm2) to the end of the bundled ground wires.
Outer covering
Attach a round crimp-on terminal
to the ground wires.
Attach a round crimp-on terminal to the end of the ground wires
(7) Strip about 7 mm of covering from the ends of each twisted-pair.
Remove about 7 mm of the covering.
Outer covering
Removing the covering from the ends of each twisted-pair
The ends of the twisted-pair cables can be screwed on to the terminal block without the covering.
To crimp a pin terminal to the ends of twisted-pairs, use the following recommended product or
equivalent.
Connection wire size:
0.75 mm2 (AWG18)
Recommended product:
Weidmüeller H 0.75/12
This completes the cable termination procedure.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
Blank Page
App.2-1
Appendix 2 Constructing Local HMI
Connection Cables
To connect a YH8000 HMI unit as a local HMI to the TDLS8000 sensor control unit (SCU), use
the special YH8000 option cable (option code: /C).
This section explains how to terminate this cable.
NOTE
• Before performing cable termination, pass the cable through cable glands. After crimping
the communication connector, the connector cannot be passed through the cable glands.
• Be careful of the cable gland orientation. The end with the screw section of the cable gland
is the connector end.
• To maintain the TDLS8000 performance and functionality, be sure to use the optional
dedicated cable.
l Required components and tools
• Wire cutter
• Pliers
l Pretreatment
Before terminating the special cable, pass the cable through cable glands (for 1/2NPT or M20).
Apply the cable glands from the unterminated end of the special cable.
After crimping the communication connector (RJ45 modular plug), the cable cannot be passed
through the cable glands.
Cable glands
Pass the cable through them
before crimping the
communication connector.
Communication connector (supplied with the dedicated cable)
Crimped
Figure 1 Local HMI connection cable
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.2-2
l Attaching the communication connector
Crimp a communication connector (RJ45 modular plug) to one end of the dedicated cable (the
end without the communication connector). Use the communication connector supplied with the
dedicated cable.
(1) Pass the wires (two pairs: orange-white and green-white) through the communication
connector case.
Pass the wires through the case.
Body
Cover
Case
Passing the wires through the communication connector case
(2) Insert the wires (two pairs: orange-white and green-white) into the communication connector
case. Separate the ends of the twisted-pairs with a nipper or similar tool, and insert each
wire in the connector case according to Table 1.
Table 1 Where to insert the wires
Cover marking
B
W
O
Y
Wire color
Green
White-green stripe
Orange
White-orange stripe
Inserting the cable wires into the connector cover
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.2-3
(3) Insert the cover with the wires into the connector body until you hear it click.
(4) Crimp the cover to the connector body using a pair of pliers or a similar tool.
Be sure not to crush the protruding parts of the connector.
Be careful not to crush the protruding parts.
Crimping the cover to the connector body
(5) After the cover has been crimped completely into the connector body, view the connector
from the side. Check that the cover is flat and that wires do not come loose from the
connector.
The cover is flat with the
connector body.
The wires do not come loose from
the connector after crimping.
Checking the crimping
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.2-4
(6) Place the case over the connector body. Check that the metallic latch on the side of the
connector engages.
Check that the metal latch is engaged to the case.
Placing the case over the connector body
(7) Insert the connector latch into the metal protruding parts of the connector body. Make sure
the metallic protruding parts at the end of the connector is not crushed.
Insert the connector latch
into the metallic protruding parts.
This completes the communication connector attachment procedure.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.3-1
Appendix 3 General View of HART DD
The entire structure of the DD menu including parameter arrangement is listed below. The menu
for a TDLS8000 with two-gas measurement specification is indicated here. Therefore, the list
includes menus and parameters that do not appear in a TDLS8000 with one-gas measurement
specification.
Note that the top menu structure is different on FieldMate. For details, see “7.2.2 DTM Menu
(FieldMate)”. Further, the labeling of some parameters is different as follows.
• Dynamic variable label names such as “PV” and “SV” are replaced with assigned item
names (e.g., “Concentration 1”).
• “PV Loop current” is displayed as “PV AO” (SV is also similar).
• “PDQ” and “LS” are displayed as “Data Quality” and “Limit Status,” respectively.
1st
Process
variables
2nd
Device variables
3rd
Dynamic
variables
Dynamic var assign
Dynamic var status
4th
5th
Item
Concentration 1
Conc 1 gas type
Concentration 2
Conc 2 gas type
Transmission
Temperature
Pressure
PV
PV Loop current
SV
SV Loop current
TV
QV
PV is
SV is
TV is
QV is
PV PDQ
PV LS
SV PDQ
SV LS
TV PDQ
TV LS
QV PDQ
QV LS
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
1st
Diagnosis/
Service
2nd
Test/Status
Logbook
Calibration
3rd
View status
Status mask
Manual
Semi-auto
Restore
Validation
Manual
Semi-auto
Transmission adjust
Loop check
Analog output
Digital output
Valve
Trim analog
channel
Trim info
System
4th
5th
App.3-2
Item
Status group 1
Status group 2
Status group 3
Status group 4
Status group 6
Status group 7
Status group 8
Status group 9
Status group 10
Device status
Ext dev status
Device Diagnostic Status 0
AO saturated
AO fixed
Cfg chng count
Reset cfg chng flag
Dev cfg locked mask
Read alarm/message record
Read cal/val record
Manual zero cal
Manual span cal
Semi-auto zero cal
Semi-auto span cal
Restore zero cal data
Restore span cal data
Clear cal alarms
Abort calibration
Manual offline val 1
Manual offline val 2
Manual online val 1
Manual online val 2
Semi-auto offline val 1
Semi-auto offline val 2
Semi-auto offline val 1+2
Semi-auto online val 1
Semi-auto online val 2
Clear val alarm
Abort validation
Transmission
Transmission adjust
Tst auto release time
AO-1 loop chk mode
AO-1 chk output
AO-2 loop chk mode
AO-2 chk output
DO-1 loop chk mode
DO-1 chk output
DO-2 (Flt) loop chk mode
DO-2 (Flt) chk output
Vlv-1 loop chk mode
Vlv-1 chk output
Vlv-2 loop chk mode
Vlv-2 chk output
Squawk
Trim AO-1 (PV)
Trim AO-2 (SV)
Trim AI-1 (Pres)
Trim AI-2 (Temp)
Trim Who
Trim Date
Trim Loc
Trim Desc
Clear latched alarms
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
1st
Basic setup
App.3-3
2nd
PV range
SV range
Assign TV & QV
3rd
4th
5th
Tag
Long tag
PV is
PV LRV
PV URV
PV unit
SV is
SV LRV
SV URV
SV unit
TV is
QV is
Item
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
1st
Detailed
setup
App.3-4
2nd
3rd
4th
I/O condition Analog output AO-1
5th
Warning hold
Fault hold
Cal/Val hold
Maintenance
hold
Warm-up
hold
AO-2
Warning hold
Fault hold
Cal/Val hold
Maintenance
hold
Warm-up
hold
Analog input AI-1 (Pressure)
AI-2 (Temperature)
Digital output DO-1 (DO)
DO-2 (Fault)
Digital input
DI-1
DI-2
Valve control
Stream time switch
Alarm
Warning
Warning group 1
Warning group 2
Item
PV is
PV LRV
PV URV
AO1 warn hld mode
AO1 warn hld level
AO1 warn hld delay
AO1 fault hld mode
AO1 fault hld level
AO1 fault hld delay
AO1 calval hld mode
AO1 calval hld level
AO1 maint hld mode
AO1 maint hld level
AO1 w-up hld mode
AO1 w-up hld level
Trim AO-1 (PV)
SV is
SV LRV
SV URV
AO2 warn hld mode
AO2 warn hld level
AO2 warn hld delay
AO2 fault hld mode
AO2 fault hld level
AO2 fault hld delay
AO2 calval hld mode
AO2 calval hld level
AO2 maint hld mode
AO2 maint hld level
AO2 w-up hld mode
AO2 w-up hld level
Trim AO-2 (SV)
Pres val at 4mA
Pres val at 20mA
Trim AI-1 (Pres)
Temp val at 4mA
Temp val at 20mA
Trim AI-2 (Temp)
DO-1 output item
DO-1 output delay
DO-2 output item
DO-2 output delay
DI-1 mode
Set DI-1 mode
DI-1 filter time
DI-2 mode
Set DI-2 mode
DI-2 filter time
Current stream
Initial stream
Stream 1 vlv pattern
Stream 2 vlv pattern
Stream 3 vlv pattern
Switch stream1 to
Stream1 duration
Switch stream2 to
Stream2 duration
Switch stream3 to
Stream3 duration
Valve usage
Set valve usage
Warn select group 1
Trans low lmt
Pres low lmt
Pres high lmt
Temp low lmt
Temp high lmt
Conc 1 low lmt
Conc 1 high lmt
Conc 2 low lmt
Warn select group 2
Conc 2 high lmt
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
1st
(Detailed
setup)
App.3-5
2nd
Calibration
3rd
Zero
calibration
4th
Valve control
Auto time
Span
calibration
Parameter
Valve control
Auto time
Zero+Span calibration
Validation
Offline
validation 1
Parameter
Valve control
Auto time
Offline
validation 2
Parameter
Valve control
Auto time
Offline validation 1+2
5th
Item
Z-cal gas purg time
Z-cal proc purg time
Z-cal auto vlv man
Z-cal time initiate
Z-cal init date
Z-cal init time
Z-cal day cycle
Z-cal hour cycle
S-cal gas type
Set s-cal gas type
S-cal gas conc
S-cal pres mode
S-cal pres fix val
S-cal temp mode
S-cal temp fix val
S-cal OPL mode
S-cal OPL fix val
S-cal gas purg time
S-cal proc purg time
S-cal auto vlv man
S-cal time initiate
S-cal init date
S-cal init time
S-cal day cycle
S-cal hour cycle
Z+S cal time initiate
Z+S cal init date
Z+S cal init time
Z+S cal day cycle
Z+S cal hour cycle
Offval1 gas type
Set offval1 gas type
Offval1 gas conc
Offval1 pres mode
Offval1 pres fix val
Offval1 temp mode
Offval1 temp fix val
Offval1 OPL mode
Offval1 OPL fix val
Offval1 gas purg time
Offval1 prc purg time
Offval1 auto vlv man
Offval1 time initiate
Offval1 init date
Offval1 init time
Offval1 day cycle
Offval1 hour cycle
Offval2 gas type
Set offval2 gas type
Offval2 gas conc
Offval2 pres mode
Offval2 pres fix val
Offval2 temp mode
Offval2 temp fix val
Offval2 OPL mode
Offval2 OPL fix val
Offval2 gas purg time
Offval2 prc purg time
Offval2 auto vlv man
Offval2 time initiate
Offval2 init date
Offval2 init time
Offval2 day cycle
Offval2 hour cycle
Offval1+2 time init
Offval1+2 init date
Offval1+2 init time
Offval1+2 day cycle
Offval1+2 hour cycle
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
1st
(Detailed
setup)
App.3-6
2nd
(Validation)
3rd
Online
validation 1
4th
Parameter
Valve control
Auto time
Conc reading mode
Online
validation 2
Parameter
Valve control
Auto time
Conc reading mode
Field device
info
Write protect menu
Memo
System
Date/time
Local display SCU LCD display
LU LED display
Communication HART output
TCP/IP
5th
Item
Onval1 gas type
Set onval1 gas type
Onval1 gas conc
Onval1 temp mode
Onval1 temp fix val
Onval1 act amb ofst
Onval1 pres fix val
Onval1 OPL fix val
Onval1 gas purg time
Onval1 nml purg time
Onval1 auto vlv man
Onval1 time initiate
Onval1 init date
Onval1 init time
Onval1 day cycle
Onval1 hour cycle
Onval1 read mode
Onval1 output factor
Onval2 gas type
Set onval2 gas type
Onval2 gas conc
Onval2 temp mode
Onval2 temp fix val
Onval2 act amb ofst
Onval2 pres fix val
Onval2 OPL fix val
Onval2 gas purg time
Onval2 nml purg time
Onval2 auto vlv man
Onval2 time initiate
Onval2 init date
Onval2 init time
Onval2 day cycle
Onval2 hour cycle
Onval2 read mode
Onval2 output factor
Descriptor
Message
Date
Final asmbly num
Write protect
Enable write 10min
New password
Software seal
Memo 1
Memo 2
Memo 3
Long tag
Present date/time
Set date/time
LCD spect disp mode
LCD backlight bright
LCD contrast
LED brightness
Poll addr
Loop current mode
Num req preams
Num resp preams
IP address
Subnet mask
Default gateway
Set IP settings
Initialize config
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
1st
(Detailed
setup)
App.3-7
2nd
Analysis
3rd
Process
param
4th
Pressure
Temperature
Non process
param
No-prcs temp
No-prcs conc
Unit
Averaging
Zero offset
5th
Item
OPL
Pres mode
Pres fix mode val
Pres active type
Pres backup mode
Pres backup set val
Temp mode
Temp fix mode val
Temp active type
Temp backup mode
Temp backup set val
Temp act amb ofst
No-prcs OPL
No-prcs pres
No-prcs temp mode
No-prcs temp fix val
No-prcs act amb coef
No-prcs conc 1
No-prcs conc 2
OPL unit
Pres unit
Temp unit
Averaging number
Averaging time
Zero offset 1
Zero offset 2
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
1st
Review
App.3-8
2nd
Process info
3rd
I/O info
I/O assignment
System info
Factory info
Field device
info
Version #’s
Additional info
Revision #'s
4th
5th
Item
Concentration 1
Conc 1 STDEV
Concentration 2
Conc 2 STDEV
Transmission
Temperature
Pressure
Conc 1 gas type
Conc 2 gas type
Temp mode
Pres mode
AI-1 (pres)
AI-2 (temp)
PV Loop current
SV Loop current
DI-1
DI-2
DO-1
DO-2 (Fault)
Valve-1
Valve-2
PV is
SV is
DI-1 mode
DI-2 mode
DO-1 output item
Long tag
IP address
Subnet mask
Default gateway
Model name
Analyzer SN
Laser module SN
SI unit control
Analysis period
Software ver
Write protect
Poll addr
Loop current mode
Num req preams
Num resp preams
Dev id
Tag
Long Tag
Descriptor
Message
Date
Distributor
Final asmbly num
Max dev vars
Device Profile
Universal rev
Fld dev rev
Software rev
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.4-1
Appendix 4 YH8000 Menu Tree
This section provides the tree structure of the setup and execution menus of the YH8000 HMI
unit.
(1) TDLS8000 Tunable Diode Laser Spectrometer operation panel
The TDLS8000 setup and execution menu tree is shown below. The menu for a TDLS8000 with
two-gas measurement specification is indicated here. Therefore, some items will not appear in a
TDLS8000 with one-gas measurement specification.
1st
Execution
=> Select analyzer => Input password =>
2nd
Calibration
3rd
Manual
4th
Tab
Zero Calibration
Span Calibration
Zero Calibration
Span Calibration
Zero + Span Calibration
Zero Calibration
Span
Gas1
Calibration
Gas2
Item
Zero Calibration
Span Calibration
Semi-Auto
Zero Calibration
Span Calibration
Zero + Span Calibration
Restore
Restore Zero Calibration data
Restore Span Calibration data
(Gas1)
Restore Span Calibration data
(Gas2)
Clear Calibration Alarm
Clear Calibration Alarm
Validation
Manual
Offline Validation 1
Offline Validation 1
Offline Validation 2
Offline Validation 2
Online Validation 1
Online Validation 1
Online Validation 2
Online Validation 2
Semi-Auto
Offline Validation 1
Offline Validation 1
Offline Validation 2
Offline Validation 2
Offline Validation 1+2
Offline Validation 1+2
Online Validation 1
Online Validation 1
Online Validation 2
Online Validation 2
Clear Validation Alarm
Clear Validation Alarm
Transmission Adjustment
Transmission Adjustment
Loop Check
Analog
AO-1 Loop check mode
AO-1 Check output
Output
AO-2 Loop check mode
AO-2 Check output
Digital Output DO Loop check mode
DO Check output
Fault Loop check mode
Fault Check output
Valve
SV-1 Loop check mode
SV-1 Check output
SV-2 Loop check mode
SV-2 Check output
Auto Release Auto release time
System
Clear Latched Alarms
Clear Latched Alarms
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
1st
2nd
Configuration I/O
3rd
Analog
Output
4th
AO-1
Tab
Output
Hold mode
AO-2
Calibration
Output
Hold mode
Analog Input AI-1
(Pressure)
Calibration
Scaling
Calibration
AI-2
Scaling
(Temperature)
Calibration
Digital Output DO-1 (DO)
Digital Input
DO-2 (Fault)
DI-1
DI-2
Valve Control
Alarm
Calibration
Zero Calibration
Valve
Auto Time
App.4-2
Item
Item
4 mA
20 mA
Warning
Delay
Fault
Delay
Cal/Val
Maintenance
Warm-up
Execute AO-1 Calibration
Item
4 mA
20 mA
Warning
Delay
Fault
Delay
Cal/Val
Maintenance
Warm-up
Execute AO-2 Calibration
4mA value
20mA value
Execute A1-1 Calibration
4mA value
20mA value
Execute A1-2 Calibration
Output item
Output delay
Output delay
Mode
Filter time
Mode
Filter time
Valve Usage
Current
Stream 1 Switch to
Stream 1 Duration
Stream 2 Switch to
Stream 2 Duration
Stream 3 Switch to
Stream 3 Duration
Initial
Transmission Low
Process Pressure Low
Process Pressure High
Process Temperature Low
Process Temperature High
Concentration Gas1 Low
Concentration Gas1 High
Concentration Gas2 Low
Concentration Gas2 High
Calibration gas Purge time
Process gas Purge time
Auto Valve for Manual Cal.
Time Initiate
Initial time
Cycle(day)
Cycle(hour)
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
1st
2nd
(Configuration) (Calibration)
3rd
Span Calibration
4th
Tab
Parameters
Valve
Auto Time
Validation
Zero + Span Calibration
Auto Time
Offline Validation 1
Parameters
Valve
Auto Time
Offline Validation 2
Parameters
Valve
Auto Time
Offline Validation 1+2
Auto Time
Online Validation 1
Parameters
Valve
Auto Time
Reading
mode
App.4-3
Item
Gas type
Concentration
Pressure
Fixed value
Temperature
Fixed value
OPL
Fixed value
Calibration gas Purge time
Process gas Purge time
Auto Valve for Manual Cal.
Time Initiate
Initial time
Cycle(day)
Cycle(hour)
Time Initiate
Initial time
Cycle(day)
Cycle(hour)
Gas type
Concentration
Pressure
Fixed value
Temperature
Fixed value
OPL
Fixed value
Validation gas Purge time
Process gas Purge time
Auto Valve for Manual Val.
Time Initiate
Initial time
Cycle(day)
Cycle(hour)
Gas type
Concentration
Pressure
Fixed value
Temperature
Fixed value
OPL
Fixed value
Validation gas Purge time
Process gas Purge time
Auto Valve for Manual Val.
Time Initiate
Initial time
Cycle(day)
Cycle(hour)
Time Initiate
Initial time
Cycle(day)
Cycle(hour)
Gas type
Concentration
Pressure
Temperature
Offset Value
Fixed Value
OPL
Validation gas Purge time
Normal gas Purge time
Auto Valve for Manual Val.
Time Initiate
Initial time
Cycle(day)
Cycle(hour)
Mode
Output Factor
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
1st
2nd
(Configuration) (Validation)
3rd
4th
Online Validation 2
Tab
Parameters
Valve
Auto Time
Reading mode
System
Tag
Date & Time
Password
Local Display
SCU
Communication TCP/IP
LU
HART
Configuration Initialization
Analysis
Process Parameters
Path Length
Pressure
Temperature
Non-Process Parameter
Path Length
Pressure
Temperature
Concentration
Units
Average
Zero Offset
App.4-4
Item
Gas type
Concentration
Pressure
Temperature
Offset Value
Fixed Value
OPL
Validation gas Purge time
Normal gas Purge time
Auto Valve for Manual Val.
Time Initiate
Initial time
Cycle(day)
Cycle(hour)
Mode
Output Factor
Tag
Current Password
New Password
Confirm New Password
Operation
Date
Time
Spectrum
Brightness
Contrast
Brightness
IP Address
Subnet Mask
Default Gateway
HART Address
Loop Current Mode
Setting data
Calibration data for AO and AI
Zero/Span Calibration data
Tag, Network setting, Password
Path Length
Mode
Active Type
Fixed Mode Value
Backup Mode
Backup Set Value
Mode
Active Type
Fixed Mode Value
Backup Mode
Backup Set Value
Offset value
Path Length
Pressure
Mode
Fixed Value
Ambient Coefficient
Gas1
Gas2
Path Length
Pressure
Temperature
Average number
Average time
Zero offset for Gas1
Zero offset for Gas2
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.4-5
(2) YH8000 HMI unit operation panel
The YH8000 HMI unit setup and execution menu tree is shown below.
=> HMI =>
1st
2nd
Analyzer Connection
Display
Setting
Home Style
Meter range
3rd
4th
Tab
Analyzer 1
Analyzer 2
Analyzer 3
Analyzer 4
Backlight
Network Setting
HMI Information
Item
Change IP
Input Analyzer IP manually
Select Analyzer by Auto-search
Connect
Disconnect
Meter Type
Gas1 Min
Gas1 Max
Gas2 Min
Gas2 Max
Gas1 Min
Gas1 Max
Gas2 Min
Gas2 Max
Gas1 Min
Gas1 Max
Gas2 Min
Gas2 Max
Gas1 Min
Gas1 Max
Gas2 Min
Gas2 Max
Brightness
Auto Off
IP Address
Subnet Mask
Default Gateway
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
Blank Page
App.5-1
Appendix 5 What Is an Analysis Period?
The TDLS8000 calculates process gas concentration from a value obtain by integrating the
spectrum data over a given period. This integration period is the analysis period.
Measured values and analog output are updated every analysis period. The analysis period is set
to an optimal value depending on the application and cannot be changed.
On the TDLS8000, you can specify how many analysis periods of spectrum data to calculate the
moving average over. The number of times moving average is taken in a single concentration
calculation is called the average number, and the corresponding time is called average time.
The average number is variable. The average time can be increased by increasing the average
number in order to reduce the influence of disturbance existent in the measurement process.
Even if the average number is increased, measured values and analog output are updated
according to the analysis period, but the analysis responsiveness declines.
The analysis period and average number are set to optimal values according to the process to be
measured before factory shipment. The average number set before factory shipment is called the
basic average number. The final average time is determined as follows.
Average time = (analysis period × basic average number) × average number)
Given a basic average number of 2, the following figures illustrate the moving average ranges
when the average number is changed.
indicates an interval during which a spectrum is acquired and the concentration is
calculated.
l When the average number is 1
(Average time = analysis period × basic average number × average number = analysis period × 2)
Analysis period
Average time
Time
Measured value and analog output updating
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.5-2
l When the average number is 2
(Average time = analysis period × 2 × 2 = analysis period × 4)
Analysis period
Average time
Time
Measured value and analog output updating
You can view the analysis period from the following menu.
[HART] “Review>>Factory info>>Analysis period”
[YH8000] “
>>System Information>>Analysis Period”
For instructions on how to set the average number and how to view the average time, see “6.9.6
Moving Average Count for Analysis Values”.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.6-1
Appendix 6 Maintaining Good
Transmission
l High dust application
90
.5
Unit: mm
Ø1
ANSI Class 150-3B 316LSS flange
305
41
44.5
48.3
Ø19
0.5
Depending on the measurement location, the process may include large amounts of
contamination, dust, and particles. In such conditions, the amount of laser beam that reaches the
SCU decreases.
Normally, such information should be conveyed to Yokogawa through the process information
questionnaire of the Application Data Sheet and assessed in the TDLS8000 design, engineering,
and specification determination stages before TDLS8000 installation. However, depending on
the process, the contamination, dust, and particle conditions may change (e.g., waste disposal
sites, thermal oxidation).
To improve transmission in a high particulate concentration process, you must reduce the
process optical path length and perform process window purge correctly and thoroughly.
RF
1-1/2 inch schedule 40
Hastelloy-C276
Ø1
27
23.4 RF to RF
5/8 inch 11 UNC
Figure 1 Example of Insertion tubes
For high dust concentration applications, please consult with Yokogawa. If it is deemed that
reducing the process optical path length can significantly improve the transmission level (or
at least to within the tolerance of the TDLS8000 capacity), we will discuss with you the special
insertion tube specification requirements. In many cases, insertion tubes are made of materials
compatible with the process medium. In applications that include acidic or corrosive gas,
Hastelloy C276 or Monel 400 is used.
l Handling process nozzle clogging problems
If dust accumulates repeatedly in process flanges or insertion tubes, try the following measures.
• Increase the flow velocity and pressure of the process window purge gas
• Use a process nozzle or insertion purge tube with small opening.
• Keep the process nozzle warm to prevent condensation or solidification
• Combine the above methods
• In the worst case, move the TDLS8000 to a cleaner measurement location (e.g.,
downstream side of ESP, knockout drum, filter scrubber).
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.6-2
Unit: mm
1610
Alignment
Flange
Thermal insulation
Laser module area
valve
1000
LU
Online
Validation cell
SCU
ANSI Class 150-3B flange
ANSI Class 150-4 B flange
40-45°C
45 to 50°C (outer temperature)
65°C (inner temperature)
150°C
Process nozzle
(insertion purge tube)
Figure 2 Installation example of insertion tubes with thermal insulation
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.7-1
Appendix 7 Safety Instrumented System
Installation
WARNING
When using the TDLS8000 as a Safety Instrumented Systems (SIS), in order to maintain the
necessary level of safety, strictly observe the instructions and procedures provided in this
Appendix.
n Scope and Purpose
This section describes the handling precautions to be taken when installing and operating the
TDLS8000 in order to maintain the level of safety designed for using the TDLS8000 in a Safety
Instrumented System application. It also provides an overview of the operation. The topics
discussed in this section are the TDLS8000’s proof test, repairs, and replacement; safety data;
service life; environmental and application limitations; and parameter settings.
n Using the TDLS8000 in a Safety Instrumented System Application
l Safety accuracy
The following table shows the TDLS8000 safety accuracy. When an error caused by an
internal component failure exceeds the safety accuracy, the TDLS8000 is considered to have
malfunctioned.
Measured gas
O2
CO(ppm)
CO+CH4
CO
CH4
NH3
H2O(ppm) in nonHC
H2O(ppm) in HC
%CO
CO+CO2
CO
CO2
NH3+H2O
NH3
H2O
%CO2_High Range.
%CO2_Extend. Range.
H2O(%)
HCl(ppm)
H2S(%)
HF(ppm)
Safety accuracy
+/-0.2% O2 or 15% of reading, whichever is greater
+/-50 ppm CO or 15% of reading, whichever is greater
+/-50 ppm CO or 15% of reading, whichever is greater
+/-0.1% CH4 or 15% of reading, whichever is greater
+/-0.1 ppm NH3 or 15% of reading, whichever is greater
+/-3 ppm H2O or 15% of reading, whichever is greater
+/-3 ppm H2O or 15% of reading, whichever is greater
+/-0.3% CO or 15% of reading, whichever is greater
+/-0.4% CO or 15% of reading, whichever is greater
+/-0.4% CO2 or 15% of reading, whichever is greater
+/-10 ppm NH3 or 15% of reading, whichever is greater
+/-1% H2O or 15% of reading, whichever is greater
+/-0.03% CO2 or 15% of reading, whichever is greater
+/-0.4% CO2 or 15% of reading, whichever is greater
+/-0.3% H2O or 15% of reading, whichever is greater
+/-3 ppm HCl or 15% of reading, whichever is greater
+/-0.6% H2S or 15% of reading, whichever is greater
+/-3 ppm HF or 15% of reading, whichever is greater
l Diagnostic response time
The TDLS8000 can indicate an internal malfunction within 30 seconds.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.7-2
l I/O restriction
Only analog output AO-1 and analog inputs AI-1 and AI-2 comply with Safety Instrumented
System. Do not use other inputs or outputs as part of a Safety Instrumented System.
l Opening and closing the TDLS8000
When online, do not open or close the cover. If you need to open and close the TDLS8000 cover
for maintenance, obtain permission from you safety administrator.
l Configuration
Use the HART Configuration tool or a YH8000 HMI unit to set the range and unit. Connect the
HART Configuration tool or the YH8000 according to the instructions in this manual.
After installing the TDLS8000, check that the range and unit are set correctly. Calibrate the
TDLS8000 after setting the parameters.
l Connecting External Transmitters
If you want to connect external transmitters for temperature or pressure input, use products
that, when used by themselves, comply with Safety Integrity Level (SIL) 2 based on a PFDavg
calculation of the entire safety instrumented function or in a redundant configuration, Safety
Integrity Level (SIL) 3 based on a PFDavg calculation of the entire safety instrumented function.
For details on installation and operation of the external transmitters in safety applications, see the
relevant safety manuals.
Temperature and pressure transmitters that we recommend are shown below.
Temperature transmitter YOGOGAWA YTA series
Pressure transmitter
YOKOGAWA EJX and EJA series
l Setting required parameters
To maintain the appropriate level of safety, set the following parameters.
Parameter
Warm-up current setting
Description
Using the HART configuration tool or the YH8000 HMI unit,
set the output of AO-1 during warm-up to Preset hold and the
output value to 3.8 mA Preset hold.
Warning-in-effect current setting Using the HART configuration tool or the YH8000 HMI unit,
set the output of AO-1 for when a warning is in effect to
Non-hold or Hold. If you need to specify Preset hold, set a
value different from the fault-in-effect current value (burnout
current) to distinguish this value from the output for when a
fault occurs.
Fault-in-effect current setting
Using the HART configuration tool or the YH8000 HMI unit,
set the output of AO-1 for when an internal fault is detected to
Preset hold and the output value to 21.0 mA or higher or to a
burnout current of 3.6 mA or less.
Hardware write protection switch Disable the HART write function.
l Using the YH8000 HMI unit
When using the YH8000 in a system, use password protection to prevent parameter settings
from being changed in modes other than offline. The safety administrator should manage the
password properly by referring to section "6.9.3 User Password Setting".
l Proof test
You must perform a proof test in order to detect faults that are not detected through selfdiagnostics but still hinder the execution of the intended safety functions of the TDLS8000.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.7-3
The proof test interval is determined by the safety calculation that is performed for each safety
instrumented function, including the TDLS8000. To maintain the safety level of the safety
instrumentation, proof tests must be performed at a frequency determined by the safety
calculation or a higher frequency.
The following tests must be performed in a proof test. The result of proof tests must be
documented, and the documents should be handled as part of the plant’s safety management. If
a fault is detected, please consult with Yokogawa.
The operator that performs proof tests on the TDLS8000 must have a thorough knowledge of
the operation of Safety Instrumented Systems, including the bypass procedure, TDLS8000
maintenance, and change procedures.
Test method
1. Bypass the safety functions, and perform appropriate measures to
prevent malfunction.
2. Use the HART Configuration tool or YH8000 to properly execute all
diagnostics and collect the results.
3. Use the loop function of the HART Configuration tool or YH8000 to
output a burn-up current, and verify that the current is at this level.
4. Use the loop function of the HART Configuration tool or YH8000
to output a burn-down current, and verify that the current is at this
level.
5. Thoroughly check for leakages and visible damages and stains.
6. Perform two-point validation over the entire operating range.
7. Release the bypass, and restore normal operation.
Required tools
Estimated result
HART
Configuration tool
or
YH8000
Proof test
coverage: 92%
l Repair and replacement
To repair the TDLS8000 while the process is online, bypass the TDLS8000. You must perform
the bypass procedure correctly. If a fault is detected, please consult with Yokogawa. TDLS8000
replacement must be performed by a trained engineer.
l Startup time
The TDLS8000 sends valid signals within 5 minutes after power-on.
l Firmware updating
For firmware updating, please consult with Yokogawa.
l Reliability data
The FMEDA (Failure Mode, Effects and Diagnostic Analysis) report that Yokogawa provides
contains failure rates and failure modes.
When used by itself, the TDLS8000 is certified for compliance with up to Safety Integrity
Level (SIL) 2 based on a PFDavg calculation of the entire safety instrumented function. The
development process of the TDLS8000 is certified for compliance with up to SIL3. When used
in a redundant configuration, it can be used at Safety Integrity Level (SIL) 3 based on a PFDavg
calculation of the entire safety instrumented function.
When used in a redundant configuration, we recommend that the common cause factors
(β-factor) for the PFD calculation of the entire safety instrumented function be set at 5%. If the
plant operator provides “common cause failure” training and a clear, detailed maintenance
procedure for preventing common cause failures, the common cause factors (β-factor) can be
set to 2%.
l Lifetime limitation
The expected lifetime of the TDLS8000 is 10 years. The reliability data in the FMEDA report is
valid to 10 years. It is assumed that the failure rates of the TDLS8000 would increase when it is
used over 10 years. Therefore, the safety integrity level based on the reliability data given in the
FMEDA report may not be attainable.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.7-4
l Environmental limitation
The environmental limitation of the TDLS8000 is defined in this manual.
l Application limitation
If the TDLS8000 is used in an application outside the limits defined in this manual, the reliability
data is void.
n Terminology and Acronyms
l Terms
Safety
Freedom from unacceptable risk of harm
Functional Safety
The ability of a system to carry out the actions necessary to achieve or to maintain a defined
safe state for the equipment, machinery, plant, and apparatus under control of the system.
Basic Safety
The equipment must be designed and manufactured such that it protects against risk of
damage to persons by electrical shock and other hazards and against resulting fire and
explosion. The protection must be effective under all conditions of the nominal operation
and under single fault condition.
Verification
• Compliance and confirmation
The demonstration for each phase of the life-cycle that the (output) deliverables of the
phase meet the objectives and requirements specified by the inputs to the phase. The
verification is usually executed by analysis, testing, or both.
• Validation
The demonstration that the safety-related system(s) or the combination of safetyrelated system(s) and external risk reduction facilities meet, in all respects, the Safety
Requirements Specification. The validation is usually executed by testing.
• Safety Assessment
The investigation to arrive at a judgment—based on evidence—of the safety achieved by
safety-related systems.
Further definitions of terms used for safety techniques and measures and the description of
safety related systems are given in IEC 61508-4.
l Acronyms
FMEDA: Failure Mode, Effects and Diagnostic Analysis
SIF: Safety Instrumented Function
SIL: Safety Integrity Level
SIS: Safety Instrumented System
SLC: Safety Lifecycle
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-1
Appendix 8 Explosion Protected Type
Instrument
In this chapter, further requirements and differences for explosion proof type instrument are
described. For explosion protected type, the description in this chapter is prior to other description
in this User’s Manual.
CAUTION
TDLS8000 and YH8000 has been tested and certified as being explosion proof. Please note
that severe restrictions apply to these instruments’s construction, installation, external wiring,
maintenance and repair. A failure to abide by these restrictions could make the instrument a
hazard to operate.
WARNING
Tag plate which is provided as an option of TDLS8000 or YH8000 shall be hung by a wire and the
wire shall be tightly bound to the alignment flange or LAO of TDLS8000, or earth terminal block of
the enclosure of YH8000, not to insulate electrically and to avoid electrostatic charging.
TDLS8000
Alignment flange
YH8000
Tag plate
Tag plate
Earth terminal
block
Tag plate
n TDLS8000
l TDLS8000-D2 (FM Approval for US)
(1) Technical data
•
Applicable standards
FM Class 3600: 2011
FM Class 3611: 2004
FM Class 3616: 2011
FM Class 3810: 2005
NEMA 250-2003
ANSI/ISA-60079-0-2013
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-2
•
•
•
ANSI/ISA-60079-15-2012
ANSI/ISA-60079-31-2013
ANSI/IEC 60529-2004 (R2011)
Ratings
Nonincendive for
Class I; Division 2; Groups A, B, C, D; T5
Dust-Ignitionproof for Class II/III; Division 1; Groups E, F, G; T5
Class I, Zone 2, AEx nA nC IIC T5
Zone 21, AEx tb IIIC T100 °C
Enclosure
Type 4X, IP66
Specific condition of use
Precautions shall be taken to minimize the risk from electrostatic discharge or propagating
brush discharge of painted parts of the enclosure.
(2) Installation
• The installation of the equipment should be carried out by the engineers or other
professional personnel of the related expertise. The installation should not be carried out by
operators or other unprofessional personnel.
• The installation should be in accordance with NEC (the National Electric Code: ANSI/NFPA70) or local electric code.
• The equipment must only be used in an area of not more than pollution degree 2, as defined
in IEC 60664-1.
• Mounting orientation of TDLS8000 is within 30 degrees relative to the horizontal direction.
• When the cable glands and/or adapters are used, they must be Ex “d”, Ex “e”, or Ex “n” type,
and must also meet the ingress protection requirement of IP54 or more. If TDLS8000 is
installed in where more than IP54 is required as ingress protection, the cable gland and/or
adapters should be meet the requirement.
• The thread type of the cable entries shall be confirmed according to the user’s manual.
• Screws of the field wiring terminals must be tightened with specified torque values as
follows:
– Sensor Control Unit: 0.5 to 0.6 Nm,
– Laser Unit: 0.4 to 0.5 Nm
• Field wiring for ethernet communication must be in accordance with IEEE 802.3 so as to
avoid overvoltage of > 119 V.
WARNING
-
-
-
-
DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE MAY BE PRESENT
POTENTIAL ELECTROSTATIC CHARGING HAZARD
FOR INSTALLATION AND SAFE USE, READ IM11Y01D01-01
INSTALL IN ACCORDANCE WITH NFM029-A91
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-3
(3) Control Drawing
Model: TDLS8000
Date: January 16, 2015
Control Drawing (US)
㻯㼘㼍㼟㼟㻌㻵㻘㻌㻰㼕㼢㼕㼟㼕㼛㼚㻌㻞㻘㻌㻳㼞㼛㼡㼜㼟㻌㻭㻘㻌㻮㻘㻌㻯㻘㻌㻰
㻯㼘㼍㼟㼟㻌㻵㻵㻘㻌㻰㼕㼢㼕㼟㼕㼛㼚㻌㻝㻘㻌㻳㼞㼛㼡㼜㼟㻌㻱㻘㻌㻲㻘㻌㻳
㻯㼘㼍㼟㼟㻌㻵㻵㻵㻘㻌㻰㼕㼢㼕㼟㼕㼛㼚㻌㻝
㻯㼘㼍㼟㼟㻌㻵㻘㻌㼆㼛㼚㼑㻌㻞㻘㻌㻳㼞㼛㼡㼜㻌㻵㻵㻯
㻯㼘㼍㼟㼟㻌㻵㻵㻘㻌㻵㻵㻵㻘㻌㼆㼛㼚㼑㻌㻞㻝㻘㻌㻳㼞㼛㼡㼜㻌㻵㻵㻵㻯
Hazardous
(Classified) Location
㼀㼑㼙㼜㼑㼞㼍㼠㼡㼞㼑㻌㻯㼘㼍㼟㼟㻦㻌㼀㻡㻘㻌㼀㻝㻜㻜℃
Class I, Division 2, Groups A, B, C, D
㻴㼍㼦㼍㼞㼐㼛㼡㼟㻌㻔㻯㼘㼍㼟㼟㼕㼒㼕㼑㼐㻕㻌㻸㼛㼏㼍㼠㼕㼛㼚㻌
Class I, Zone 2, Group IIC
Temperature Class: T5
Unclassified Location
㼁㼚㼏㼘㼍㼟㼟㼕㼒㼕㼑㼐㻌㻸㼛㼏㼍㼠㼕㼛㼚㻌㼛㼞
㻴㼍㼦㼍㼞㼐㼛㼡㼟㻌㻔㻯㼘㼍㼟㼟㼕㼒㼕㼑㼐㻕㻌㻸㼛㼏㼍㼠㼕㼛㼚㻌
YH8000 HMI UNIT
(not necessarily be
connected)
TDLS8000 Tunable Diode
Laser Spectrometer
(Sensor Control Unit)
Power Supply
TDLS8000 Tunable Diode
Laser Spectrometer
(Laser Unit)
Control Equipment
Class I, Division 2, Groups A, B, C, D
Class II, Division 1, Groups E, F, G
Class III, Division 1
Class I, Zone 2, Group IIC
Temperature Class: T5
Zone 21
Maximum Surface Temperature: 100 °C
Rev.
Doc. No.:
NFM029-A91 P.1
Yokogawa Electric Corporation
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-4
Model: TDLS8000
Date: January 16, 2015
Specific Condition of Use:
Precautions shall be taken to minimize the risk from electrostatic discharge
or
propagating brush discharge of painted parts of the enclosure.
Notes:
1.
No revision to this drawing without prior approval of FM.
2.
Installation must be in accordance with the National Electric Code (NFPA 70) and
relevant local codes.
3.
The thread type of the cable entries must be confirmed according to the user ’s manual.
4.
Screws of the field wiring terminals must be tightened with specified torque values as
follows:
– Sensor Control Unit: 0.5 to 0.6 Nm,
– Laser Unit: 0.4 to 0.5 Nm
5.
Field wiring for ethernet communication must be in accordance with IEEE 802.3 so as
to avoid overvoltage of > 119 V.
6.
WARNING – POTENTIAL ELECTROSTATIC CHARGING HAZARD – SEE USER’S
MANUAL
7.
WARNING – DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE MAY BE
PRESENT
8.
WARNING – USE HEAT-RESISTING CABLES ≥ 70 °C
9.
WARNING – EXPOSURE TO SOME CHEMICALS MAY DEGRADE THE SEALING
PROPERTIES OF MATERIALS USED IN THE FOLLOWING RELAY INCORPORATED
IN THE SENSOR CONTROL UNIT:
ATX206SA (Manufacturer: Panasonic Corporation)
10. WARNING – SUBSTITUTION OF COMPONENTS MAY IMPAIR SUITABILITY FOR
DIVISION 2
Rev.
Doc. No.:
NFM029-A91 P.2
Yokogawa Electric Corporation
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-5
(4) Operation
WARNING
• Electrostatic charge may cause an explosion hazard. Avoid any actions that cause the
generation of electrostatic charge, such as rubbing with a dry cloth on coating face of
product.
• Take care not to generate mechanical sparking when access to the instrument and
peripheral devices in hazardous locations.
(5) Maintenance and Repair
WARNING
The instrument modification or parts replacement by other than authorized Representative of
Yokogawa Electric Corporation is prohibited and will void the certification.
When opening the cover, the enclosure should be dry and clean to prevent from ingress water or
dust.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-6
l TDLS8000-C2 (FM Approval for Canada)
Model: TDLS8000
Date: January 16, 2015
Instructions (Canada)
1) Technical data
• Applicable standards
CAN/CSA-C22.2 No. 0-10 (R2015)
CAN/CSA-C22.2 No. 25-1966 (R2014)
CAN/CSA-C22.2 No. 94.1-07 (R2012)
CAN/CSA-C22.2 No. 94.2-07 (R2012)
CAN/CSA-C22.2 No. 60079-0:11
CAN/CSA-C22.2 No. 60079-15:12
CAN/CSA C22.2 No. 60529-05 (R2010)
CAN/CSA-C22.2 No. 61010-1-12
CAN/CSA-C22.2 No. 61010-2-030-12
ANSI/ISA-12.27.01-2011
• Ratings
Ex nA nC IIC T5
Class II/III; Division 1; Groups E, F, G
• Enclosure
Type 4X, IP66
• Specific condition of use
Precautions shall be taken to minimize the risk from electrostatic discharge or
propagating brush discharge of painted parts of the enclosure.
2) Name Plate
Main Name Plate (Sensor Control Unit side)
Ex marking
Warning
(Avertissement)
Reference to
the control drawing
number
Sub Name Plate (Laser Unit side)
Warning
(Avertissement)
Reference to
the control drawing
number
MODEL:
AMB.TEMP:
SUFFIX:
OUTPUT:
NO.:
STYLE:
SUPPLY:
MANUFACTURED:
Specified model code
Specified ambient temperature range
Specified suffix code
Specified analog output range
Serial number
Specified style code
Specified supply voltage and wattage
Month and year of production
Rev.1: December 3, 2015
Doc. No.:
NFM029-A92 P.1
Yokogawa Electric Corporation
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-7
Model: TDLS8000
Ex marking:
Warning:
Avertissement:
Date: January 16, 2015
Ex nA nC IIC T5
CL II/III, DIV 1, GP EFG
TYPE 4X, IP66
DUAL SEAL
USE THE HEAT-RESISTING CABLES ≥ 70 °C
DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE MAY BE
PRESENT.
POTENTIAL ELECTROSTATIC CHARGING HAZARD.
FOR INSTALLATION AND SAFE USE, READ IM mmmm
CÂBLES RESISTANTS A UNE TEMPÉRATURE ≥70°C
NE PAS OUVRIR EN ATMOSPHÈRE EXPLOSIVE.
DANGER : ELECTRICITÉ STATIQUE.
LISEZ IM mmmm POUR INSTALLATION ET SÉCURITÉ.
Note: “mmmm” means document number of user ’s manual.
3) Installation
•
•
•
•
•
•
•
The installation of the equipment should be carried out by the engineers or other professional
personnel of the related expertise. The installation should not be carried out by operators or other
unprofessional personnel.
The installation must be in accordance with Canadian Electrical Code Part I (C22.1) and relevant
local codes.
The equipment must only be used in an area of not more than pollution degree 2, as defined in IEC
60664-1.
Mounting orientation of TDLS8000 is within 30 degrees relative to the horizontal direction.
When the cable glands and/or adapters are used, they must be Ex “d”, Ex “e”, or Ex “n” type, and
must also meet the ingress protection requirement of IP54 or more.
The thread type of the cable entries shall be confirmed according to the user’s manual.
Screws of the field wiring terminals must be tightened with specified torque values as
follows:
– Sensor Control Unit: 0.5 to 0.6 Nm,
•
– Laser Unit: 0.4 to 0.5 Nm
Field wiring for ethernet communication must be in accordance with IEEE 802.3 so as
to avoid overvoltage of > 119 V.
WARNING
--------------------------------------------------------------------------------------------------------------------------------------USE THE HEAT-RESISTING CABLES ≥ 70 °C
DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE MAY BE PRESENT.
POTENTIAL ELECTROSTATIC CHARGING HAZARD.
FOR INSTALLATION AND SAFE USE, READ IM mmmm
CÂBLES RESISTANTS A UNE TEMPÉRATURE ≥70°C
NE PAS OUVRIR EN ATMOSPHÈRE EXPLOSIVE.
DANGER : ELECTRICITÉ STATIQUE.
LISEZ IM mmmm POUR INSTALLATION ET SÉCURITÉ.
--------------------------------------------------------------------------------------------------------------------------------------------Note: “mmmm” = document number of user ’s manual.
Rev.
Doc. No.:
NFM029-A92 P.2
Yokogawa Electric Corporation
Note: If TDLS8000 is installed in where more than IP54 is required as ingress protection, the cable gland and/or adapters should be
meet the requirement.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-8
Model: TDLS8000
Date: January 16, 2015
4) Operation
WARNING
--------------------------------------------------------------------------------------------------------------------------------------------• Electrostatic charge may cause an explosion hazard. Avoid any actions that cause the generation of
electrostatic charge, such as rubbing with a dry cloth on coating face of product.
• Take care not to generate mechanical sparking when access to the instrument and peripheral
devices in hazardous locations.
--------------------------------------------------------------------------------------------------------------------------------------------5) Maintenance and Repair
WARNING
--------------------------------------------------------------------------------------------------------------------------------------------• The instrument modification or parts replacement by other than authorized Representative of
Yokogawa Electric Corporation is prohibited and will void the certification.
--------------------------------------------------------------------------------------------------------------------------------------------• When opening the cover, the enclosure should be dry and clean to prevent from ingress water or
dust.
Rev.
Doc. No.:
NFM029-A92 P.3
Yokogawa Electric Corporation
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-9
l TDLS8000-S2 (ATEX Certification)
(1) Technical Data
• Applicable standards
EN 60079-0: 2012+A11:2013
EN 60079-15: 2010
EN 60079-28: 2007
EN 60079-28: 2015
EN 60079-31: 2014
• Ratings
II 3(1) G Ex nA nC [op is T6 Ga] IIC T5 Gc
II 2 D Ex tb IIIC T100 °C Db
•
•
Enclosure
IP66 (In accordance with EN 60529)
Specific condition of use
Precautions shall be taken to minimize the risk from electrostatic discharge or propagating
brush discharge of painted parts of the enclosure.
In case of the enclosure of TDLS8000 with paint layers, if it is mounted in an area where
the use of category 2D equipment is required, it shall be installed in such a way that the risk
from electrostatic discharges and propagating brush discharges caused by rapid flow of
dust is avoided.
(2) Name Plate
MODEL:
Specified model code
AMB.TEMP:
Specified ambient temperature range
SUFFIX:
Specified suffix code
OUTPUT:
Specified analog output range
NO.:
Serial number
STYLE:
Specified style code
SUPPLY:
Specified supply voltage and wattage
MANUFACTURED: Month and year of production
Main Name Plate (Sensor Control Unit side)
Sub Name Plate (Laser Unit side)
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-10
(3) Installation
• The installation of the equipment should be carried out by the engineers or other
professional personnel of the related expertise. The installation should not be carried out by
operators or other unprofessional personnel.
• The installation must be in accordance with EN 60079-14 and relevant local codes.
• The equipment must only be used in an area of not more than pollution degree 2, as defined
in IEC 60664-1.
• Mounting orientation of TDLS8000 is within 30 degrees relative to the horizontal direction.
• When the cable glands and/or adapters are used, they must be Ex “d”, Ex “e”, or Ex “n” type,
and must also meet the ingress protection requirement of IP54 or more. If TDLS8000 is
installed in where more than IP54 is required as ingress protection, the cable gland and/or
adapters should be meet the requirement.
• The thread type of the cable entries shall be confirmed according to the user’s manual.
• Screws of the field wiring terminals must be tightened with specified torque values as
follows:
– Sensor Control Unit: 0.5 to 0.6 Nm,
– Laser Unit: 0.4 to 0.5 Nm
• Field wiring for ethernet communication must be in accordance with IEEE 802.3 so as to
avoid overvoltage of > 119 V.
WARNING
- DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE MAY BE PRESENT
- POTENTIAL ELECTROSTATIC CHARGING HAZARD
- FOR INSTALLATION AND SAFE USE, READ IM11Y01D01-01
(4) Operation
WARNING
• Electrostatic charge may cause an explosion hazard. Avoid any actions that cause the
generation of electrostatic charge, such as rubbing with a dry cloth on coating face of
product.
• Take care not to generate mechanical sparking when access to the instrument and
peripheral devices in hazardous locations.
(5) Maintenance and Repair
WARNING
The instrument modification or parts replacement by other than authorized Representative of
Yokogawa Electric Corporation is prohibited and will void the certification.
When opening the cover, the enclosure should be dry and clean to prevent from ingress water or
dust.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-11
l TDLS8000-E2 (IECEx Certification)
(1) Technical Data
•
•
•
•
Applicable standards
IEC 60079-0: 2011
IEC 60079-15: 2010
IEC 60079-28: 2015
IEC 60079-31: 2013
Ratings
Ex nA nC [op is T6 Ga] IIC T5 Gc
Ex tb IIIC T100 °C Db
Enclosure
IP66 (In accordance with IEC 60529)
Specific condition of use
Precautions shall be taken to minimize the risk from electrostatic discharge of painted parts
of the enclosure.
In case of the enclosure of TDLS8000 with paint layers, if it is mounted in an area where the
use of EPL Db equipment is required, it shall be installed in such a way that the risk from
electrostatic discharges and propagating brush discharges caused by rapid flow of dust is
avoided.
(2) Installation
• The installation of the equipment should be carried out by the engineers or other
professional personnel of the related expertise. The installation should not be carried out by
operators or other unprofessional personnel.
• The installation must be in accordance with IEC 60079-14 and relevant local codes.
• The equipment must only be used in an area of not more than pollution degree 2, as defined
in IEC 60664-1.
• Mounting orientation of TDLS8000 is within 30 degrees relative to the horizontal direction.
• When the cable glands and/or adapters are used, they must be Ex “d”, Ex “e”, or Ex “n” type,
and must also meet the ingress protection requirement of IP54 or more. If TDLS8000 is
installed in where more than IP54 is required as ingress protection, the cable gland and/or
adapters should be meet the requirement.
• The thread type of the cable entries shall be confirmed according to the user’s manual.
• Screws of the field wiring terminals must be tightened with specified torque values as
follows:
– Sensor Control Unit: 0.5 to 0.6 Nm,
– Laser Unit: 0.4 to 0.5 Nm
• Field wiring for ethernet communication must be in accordance with IEEE 802.3 so as to
avoid overvoltage of > 119 V.
WARNING
DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE MAY BE PRESENT
POTENTIAL ELECTROSTATIC CHARGING HAZARD
FOR INSTALLATION AND SAFE USE, READ IM11Y01D01-01
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-12
(3) Operation
WARNING
• Electrostatic charge may cause an explosion hazard. Avoid any actions that cause the
generation of electrostatic charge, such as rubbing with a dry cloth on coating face of
product.
• Take care not to generate mechanical sparking when access to the instrument and
peripheral devices in hazardous locations.
(4) Maintenance and Repair
WARNING
The instrument modification or parts replacement by other than authorized Representative of
Yokogawa Electric Corporation is prohibited and will void the certification.
When opening the cover, the enclosure should be dry and clean to prevent from ingress water or
dust.
l TDLS8000-D1 (FM Approval for US)
(1) Technical data
•
•
•
•
Applicable standards
Class 3600: 2011
Class 3615: 2006
Class 3616: 2011
Class 3810: 2005
NEMA 250-2003
ANSI/ISA-60079-0-2013
ANSI/ISA-60079-1-2009 (R2013)
ANSI/ISA-60079-31-2013
ANSI/IEC 60529-2004 (R2011)
Ratings:
Explosionproof for
Class I; Division 1; Groups A, B, C, D; T5
Dust-Ignitionproof for Class II/III; Division 1; Groups E, F, G; T5
Class I, Zone 1, AEx d IIC T5
Zone 21, AEx tb IIIC T100 °C
Enclosure: Type 4X, IP66
Specific condition of use:
- Flamepath joints are not for repair. Contact the manufacturer.
- Precautions shall be taken to minimize the risk from electrostatic discharge or
propagating brush discharge of painted parts of the enclosure.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-13
(2) Name Plate
MODEL:
Specified model code
AMB.TEMP:
Specified ambient temperature range
SUFFIX:
Specified suffix code
OUTPUT:
Specified analog output range
NO.:
Serial number
STYLE:
Specified style code
SUPPLY:
Specified supply voltage and wattage
MANUFACTURED: Month and year of production
Ex marking: CL I, DIV 1, GP ABCD, T5
CL II/III, DIV 1, GP EFG, T5
CL I, ZN 1, AEx d IIC T5
ZN 21 AEx tb IIIC T100°C
TYPE 4X, IP66
Main Name Plate (Sensor Control Unit side)
Sub Name Plate (Laser Unit side)
(3) Installation
• The installation of the equipment should be carried out by the engineers or other
professional personnel of the related expertise. The installation should not be carried out by
operators or other unprofessional personnel.
• The installation should be in accordance with NEC (the National Electric Code: ANSI/NFPA70) or local electric code.
• In hazardous location, wiring shall be in conduit.
WARNING
SEAL ALL CONDUITS WITHIN 18 INCHES. WHEN IN INSTALLED IN CL I, DIV1 .
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-14
Note: Laser unit shall have conduit and sealing fitting in the same manner as Sensor control unit.
• Unused apertures shall be closed with suitable flameproof certified blanking elements. The
plugs that are installed in the instrument are certified as a part of the instrument.
• Mounting orientation of TDLS8000 is within 30 degrees relative to the horizontal direction.
• If the TDLS8000 is mounted in an area where the use of Class II and III equipment is
required, it shall be installed in such a way that the risk from electrostatic discharges and
propagating burst discharges caused by rapid flow of dust is avoided.
• A mark indicating the electrical connection type is stamped near the electrical connection
port. These marks are as followed.
Screw size
ISO M20x1.5 female
M25x1.5 female
ANSI 1/2 NPT Female
3/4 NPT female
M
Location of the marking
Marking
M
N
Note: This type is only NPT thread available.
(4) Operation
• Note a warning label worded as follows.
WARNING
•
•
•
•
DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE MAY BE PRESENT.
POTENTIAL ELECTROSTATIC CHARGING HAZARD.
FOR INSTALLATION AND SAFE USE, READ IM 11Y01D01-01.
SEAL ALL CONDUITS WITHIN 18 INCHES. WHEN IN INSTALLED IN CL I, DIV1 .
• Take care not to generate mechanical spark when access to the instrument and peripheral
devices in hazardous locations.
• Electrostatic charge may cause an explosion hazard. Avoid any actions that cause the
generation of electrostatic charge, such as rubbing with dry cloth on coating face of product.
• Purge inside of the each enclosure using nitrogen for at least 60 minutes at 10 kPa before
turning on the power to prevent from pressurizing internal gas by purge gas. After stoppage
of purge gas, explosive gas may be inside of the enclosures.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
Purged compartment of
Sensor control unit (SCU Ex area)
Vent
App.8-15
Purged compartment of
Laser unit (LU Ex area)
Vent
P
P
N2 gas
(5) Maintenance
• The instrument modification or parts replacement by other than authorized representative of
Yokogawa Electric Corporation is prohibited and will void the approval.
l TDLS8000-C1 (FM Approval for Canada)
(1) Technical data
• Applicable standards
CAN/CSA-C22.2 No. 04-04 (R2013)
CAN/CSA-C22.2 No. 0.5-1982 (R2012)
C22.2 No. 25-1966 (R2014)
C22.2 No. 94.2-15
CAN/CSA-C22.2 No.60079-0: 11
CAN/CSA-C22.2 No.60079-1: 11
CAN/CSA-C22.2 No.60079-31: 12
CAN/CSA C22.2 No.60529-05 (R2010)
CAN/CSA C22.2 No.61010-1-12
ANSI/ISA-12.27.01-2011
• Ratings:
Ex d IIC T5 Gb
Class II/III; Division 1; Groups E, F, G T5
• Enclosure: Type 4X, IP66
• Ambient temperature: -20 to +55 °C
• Specific condition of use:
- Flamepath joints are not for repair. Contact the manufacturer.
- Precautions shall be taken to minimize the risk from electrostatic discharge or
propagating brush discharge of painted parts of the enclosure.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-16
(2) Name Plate
MODEL:
Specified model code
AMB.TEMP:
Specified ambient temperature range
SUFFIX:
Specified suffix code
OUTPUT:
Specified analog output range
NO.:
Serial number
STYLE:
Specified style code
SUPPLY:
Specified supply voltage and wattage
MANUFACTURED: Month and year of production
Ex marking: Ex d IIC T5
CL II/III, DIV 1, GPS EFG
TYPE 4X, IP66
DUAL SEAL
Main Name Plate (Sensor Control Unit side)
Sub Name Plate (Laser Unit side)
(3) Installation
• The installation of the equipment should be carried out by the engineers or other
professional personnel of the related expertise. The installation should not be carried out by
operators or other unprofessional personnel.
• All wiring shall comply with Canadian Electrical Code Part I and Local Electrical Codes.
• In hazardous location, wiring shall be in conduit.
WARNING
AVERTISSEMENT
A SEAL SHALL BE INSTALLED WITHIN 50 mm OF THE ENCLOSURE.
PLACER UN MATÉRIAU ÉTANCHE A 50 mm AUTOUR DU BOÎTIER.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-17
50 mm
50 mm
Note: Laser unit shall have conduit and sealing fitting in the same manner as Sensor control unit.
• Unused apertures shall be closed with suitable flameproof certified blanking elements. The
plugs that are installed in the instrument are certified as a part of the instrument.
• In case of ANSI 1/2 and 3/4 plugs, ANSI hexagonal wrench should be applied to screw in.
• Mounting orientation of TDLS8000 is within 30 degrees relative to the horizontal direction.
• If the TDLS8000 is mounted in an area where the use of Class II and III equipment is
required, it shall be installed in such a way that the risk from electrostatic discharges and
propagating burst discharges caused by rapid flow of dust is avoided.
• A mark indicating the electrical connection type is stamped near the electrical connection
port. These marks are as followed.
Screw size
ISO M20x1.5 female
M25x1.5 female
ANSI 1/2 NPT Female
3/4 NPT female
M
Marking
M
N
Note: This type is only NPT thread available.
Location of the marking
(4) Operation
• Note a warning label worded as follows.
WARNING
•
•
•
•
•
•
•
•
AVERTISSEMENT
DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE MAY BE PRESENT.
POTENTIAL ELECTROSTATIC CHARGING HAZARD.
FOR INSTALLATION AND SAFE USE, READ IM 11Y01D01-01.
A SEAL SHALL BE INSTALLED WITHIN 50 mm OF THE ENCLOSURE.
NE PAS OUVRIR NE ATMOSOHERE EXPLOSIVE.
DANGER : ELECTRICITÉ STATIQUE
LISEZ IM 11Y01D01-01 POUR INSTALLATION ET SÉCURITÉ.
PLACER UN MATÉRIAU ÉTANCHE A 50 mm AUTOUR DU BOÎTIER.
• Take care not to generate mechanical spark when access to the instrument and peripheral
devices in hazardous locations.
• Electrostatic charge may cause an explosion hazard. Avoid any actions that cause the
generation of electrostatic charge, such as rubbing with dry cloth on coating face of product.
• Purge inside of the each enclosure using nitrogen for at least 60 minutes at 10 kPa before
turning on the power to prevent from pressurizing internal gas by purge gas. After stoppage
of purge gas, explosive gas may be inside of the enclosures.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
Purged compartment of
Sensor control unit (SCU Ex area)
Vent
App.8-18
Purged compartment of
Laser unit (LU Ex area)
Vent
P
P
N2 gas
(5) Maintenance
• The instrument modification or parts replacement by other than authorized representative of
Yokogawa Electric Corporation is prohibited and will void the approval.
l TDLS8000-S1 (ATEX certification)
(1) Technical Data
• Applicable standards
EN 60079-0: 2012/A12:2013
EN 60079-1: 2007
EN 60079-28: 2007
EN 60079-28: 2015
EN 60079-31: 2014
• Ratings
II 2(1) G Ex d [op is T6 Ga] IIC T5 Gb
II 2 D Ex tb IIIC T100 °C Db
• Enclosure: IP65
• Ambient temperature:
-20 to +55 °C
• Specific condition of use
- Flamepath joints are not for repair. Contact the manufacturer.
- Precautions shall be taken to minimize the risk from electrostatic discharge or
propagating brush discharge of painted parts of the enclosure.
- In case of the enclosure of TDLS8000 with paint layers, if it is mounted in an area where
the use of category 2D equipment is required, it shall be installed in such a way that the
risk from electrostatic discharges and propagating brush discharges caused by rapid flow
of dust is avoided.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-19
(2) Name Plate
MODEL:
Specified model code
AMB.TEMP:
Specified ambient temperature range
SUFFIX:
Specified suffix code
OUTPUT:
Specified analog output range
NO.:
Serial number
STYLE:
Specified style code
SUPPLY:
Specified supply voltage and wattage
MANUFACTURED: Month and year of production
Ex marking: Ex d [op is T6 Ga] IIC T5 Gb
Ex tb IIIC T100 °C Db
IP66
No. FM15ATEX0042 X
Main Name Plate (Sensor Control Unit side)
Sub Name Plate (Laser Unit side)
(3) Installation
• The installation of the equipment should be carried out by the engineers or other
professional personnel of the related expertise. The installation should not be carried out by
operators or other unprofessional personnel.
• All wiring shall comply with local installation requirement.
• Cable glands, adapters and/or blanking element with a suitable IP rating shall be of Ex d IIC/
Ex tb IIIC certified by ATEX and shall be installed so as to maintain the specific degree of
protection of the equipment.
• Unused apertures shall be closed with suitable flameproof certified blanking elements. The
plugs that are installed in the instrument are certified as a part of the instrument.
• In case of ANSI 1/2 and 3/4 plugs, ANSI hexagonal wrench should be applied to screw in.
• Mounting orientation of TDLS8000 is within 30 degrees relative to the horizontal direction.
• If the TDLS8000 is mounted in an area where the use of EPL Db equipment is required, it
shall be installed in such a way that the risk from electrostatic discharges and propagating
burst discharges caused by rapid flow of dust is avoided.
• A mark indicating the electrical connection type is stamped near the electrical connection
port. These marks are as followed.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-20
Screw size
ISO M20x1.5 female
M25x1.5 female
ANSI 1/2 NPT Female
3/4 NPT female
M
Marking
M
N
Location of the marking
(4) Operation
• Note a warning label worded as follows.
WARNING
• DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE MAY BE PRESENT.
• POTENTIAL ELECTROSTATIC CHARGING HAZARD.
• FOR INSTALLATION AND SAFE USE, READ IM 11Y01D01-01.
• Take care not to generate mechanical spark when access to the instrument and peripheral
devices in hazardous locations.
• Electrostatic charge may cause an explosion hazard. Avoid any actions that cause the
generation of electrostatic charge, such as rubbing with dry cloth on coating face of product.
• Purge inside of the each enclosure using nitrogen for at least 60 minutes at 10 kPa before
turning on the power to prevent from pressurizing internal gas by purge gas. After stoppage
of purge gas, explosive gas may be inside of the enclosures.
Purged compartment of
Sensor control unit (SCU Ex area)
Vent
Purged compartment of
Laser unit (LU Ex area)
Vent
P
P
N2 gas
(5) Maintenance
• The instrument modification or parts replacement by other than authorized representative of
Yokogawa Electric Corporation is prohibited and will void the approval.
l TDLS8000-E1 (IECEx Certification)
(1) Technical Data
•
Applicable standards
IEC 60079-0: 2011
IEC 60079-1: 2007
IEC 60079-28: 2015
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-21
IEC 60079-31: 2013
• Ratings
Ex d [op is T6 Ga] IIC T5 Gb
Ex tb IIIC T100 °C Db
• Enclosure: IP66
• Ambient temperature:
-20 to +55 °C
• Specific condition of use
- Flamepath joints are not for repair. Contact the manufacturer.
- Precautions shall be taken to minimize the risk from electrostatic discharge of painted
parts of the enclosure.
- In case of the enclosure of TDLS8000 with paint layers, if it is mounted in an area where
the use of EPL Db equipment is required, it shall be installed in such a way that the risk
from electrostatic discharges and propagating brush discharges caused by rapid flow of
dust is avoided.
(2) Name Plate
MODEL:
Specified model code
AMB.TEMP:
Specified ambient temperature range
SUFFIX:
Specified suffix code
OUTPUT:
Specified analog output range
NO.:
Serial number
STYLE:
Specified style code
SUPPLY:
Specified supply voltage and wattage
MANUFACTURED: Month and year of production
Ex marking: Ex d [op is T6 Ga] IIC T5 Gb
Ex tb IIIC T100 °C Db
IP66
No. IECEx FMG 15.0024 X
Main Name Plate (Sensor Control Unit side)
Sub Name Plate (Laser Unit side)
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-22
(3) Installation
• The installation of the equipment should be carried out by the engineers or other
professional personnel of the related expertise. The installation should not be carried out by
operators or other unprofessional personnel.
• All wiring shall comply with local installation requirement.
• Cable glands, adapters and/or blanking element with a suitable IP rating shall be of Ex d IIC/
Ex tb IIIC certified by IECEx and shall be installed so as to maintain the specific degree of
protection of the equipment.
• Unused apertures shall be closed with suitable flameproof certified blanking elements. The
plugs that are installed in the instrument are certified as a part of the instrument.
• In case of ANSI 1/2 and 3/4 plugs, ANSI hexagonal wrench should be applied to screw in.
• Mounting orientation of TDLS8000 is within 30 degrees relative to the horizontal direction.
• If the TDLS8000 is mounted in an area where the use of EPL Db equipment is required, it
shall be installed in such a way that the risk from electrostatic discharges and propagating
burst discharges caused by rapid flow of dust is avoided.
• A mark indicating the electrical connection type is stamped near the electrical connection
port. These marks are as followed.
M
Location of the marking
Screw size
ISO M20x1.5 female
M25x1.5 female
ANSI 1/2 NPT Female
3/4 NPT female
Marking
M
N
(4) Operation
• Note a warning label worded as follows.
WARNING
• DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE MAY BE PRESENT.
• POTENTIAL ELECTROSTATIC CHARGING HAZARD.
• FOR INSTALLATION AND SAFE USE, READ IM 11Y01D01-01.
• Take care not to generate mechanical spark when access to the instrument and peripheral
devices in hazardous locations.
• Electrostatic charge may cause an explosion hazard. Avoid any actions that cause the
generation of electrostatic charge, such as rubbing with dry cloth on coating face of product.
• Purge inside of the each enclosure using nitrogen for at least 60 minutes at 10 kPa before
turning on the power to prevent from pressurizing internal gas by purge gas. After stoppage
of purge gas, explosive gas may be inside of the enclosures.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
Purged compartment of
Sensor control unit (SCU Ex area)
Vent
App.8-23
Purged compartment of
Laser unit (LU Ex area)
Vent
P
P
N2 gas
(5) Maintenance
• The instrument modification or parts replacement by other than authorized representative of
Yokogawa Electric Corporation is prohibited and will void the approval.
n YH8000
l YH8000-D2 (FM Approval for US)
(1) Technical Data
• Applicable standards
FM Class 3600: 2011
FM Class 3611: 2004
FM Class 3810: 2005
NEMA 250-2003
ANSI/ISA-60079-0-2013
ANSI/ISA-60079-11-2014
ANSI/ISA-60079-15-2012
ANSI/IEC 60529-2004 (R2011)
• Ratings
Nonincendive for
Class I; Division 2; Groups A, B, C, D; T5
Class I, Zone 2, AEx nA ic IIC T5
• Enclosure
Type 4X, IP65
• Specific condition of use
Precautions shall be taken to minimize the risk from electrostatic discharge of painted parts
and non-metallic parts of the enclosure.
(2) Installation
• The installation of the equipment should be carried out by the engineers or other
professional personnel of the related expertise. The installation should not be carried out by
operators or other unprofessional personnel.
• The installation must be in accordance the National Electric Code (NFPA 70) and relevant
local codes.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-24
• The equipment must only be used in an area of not more than pollution degree 2, as defined
in IEC 60664-1.
• When the cable glands and/or adapters are used, they must be Ex “d”, Ex “e”, or Ex “n”
type, and must also meet the ingress protection requirement of IP54 or more. If YH8000 is
installed in where more than IP54 is required as ingress protection, the cable gland and/or
adapters should be meet the requirement.
• The thread type of the cable entries shall be confirmed according to the user’s manual.
• Screws of the field wiring terminals must be tightened with the specified torque value: 0.22
to 0.25 Nm.
• Field wiring for ethernet communication must be in accordance with IEEE 802.3 so as to
avoid overvoltage of > 119 V.
WARNING
- DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE MAY BE PRESENT
- POTENTIAL ELECTROSTATIC CHARGING HAZARD
- INSTALL IN ACCORDANCE WITH NFM030-A71
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-25
(3) Control Drawing
Model: YH8000
Date: January 23, 2015
Control Drawing (US)
Unclassified Location
Hazardous (Classified) Location
Class I, Division 2, Groups A, B, C, D
Class I, Zone 2, Group IIC
Temperature Class: T5
YH8000 HMI UNIT
Note 1
Note 2
TDLS8000 Tunable Diode
Laser Spectrometer
(Sensor Control Unit)
Power Supply
TDLS8000 Tunable Diode
Laser Spectrometer
(Laser Unit)
Control Equipment
Class I, Division 2, Groups A, B, C, D
Class II, Division 1, Groups E, F, G
Class III, Division 1
Class I, Zone 2, Group IIC
Temperature Class: T5
Zone 21
Maximum Surface Temperature: 100 °C
Note 1: Connection only for local HMI configuration
Note 2: Connection only for remote HMI configuration
Rev.
Doc. No.:
NFM030-A71 P.1
Yokogawa Electric Corporation
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-26
Model: YH8000
Date: January 23, 2015
Specific Condition of Use:
-Precautions shall be taken to minimize the risk from electrostatic discharge of painted
parts and non-metallic parts of the enclosure.
Notes:
1.
No revision to this drawing without prior approval of FM.
2.
Installation must be in accordance with the National Electric Code (NFPA 70) and
relevant local codes.
3.
The thread type of the cable entries must be confirmed according to the user ’s manual.
4.
Screws of the field wiring terminals must be tightened with the specified torque value:
0.22 to 0.25 Nm.
5.
Field wiring for ethernet communication must be in accordance with IEEE 802.3 so as
to avoid overvoltage of > 119 V.
6.
WARNING – POTENTIAL ELECTROSTATIC CHARGING HAZARD – SEE USER’S
MANUAL
7.
WARNING – DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE MAY BE
PRESENT
8.
WARNING – SUBSTITUTION OF COMPONENTS MAY IMPAIR SUITABILITY FOR
DIVISION 2
Rev.1: December 24, 2015
Doc. No.:
NFM030-A71 P.2
Yokogawa Electric Corporation
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-27
(4) Operation
WARNING
• Electrostatic charge may cause an explosion hazard. Avoid any actions that cause the
generation of electrostatic charge, such as rubbing with a dry cloth on coating face of
product.
• Take care not to generate mechanical sparking when access to the instrument and
peripheral devices in hazardous locations.
(5) Maintenance and Repair
WARNING
The instrument modification or parts replacement by other than authorized Representative of
Yokogawa Electric Corporation is prohibited and will void the certification.
When opening the cover, the enclosure should be dry and clean to prevent from ingress water or
dust.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-28
l YH8000-C2 (FM Approval for Canada)
Model: YH8000
Date: January 23, 2015
Instructions (Canada)
1) Technical Data
• Applicable standards
CAN/CSA-C22.2 No. 0-10 (R2015)
CAN/CSA-C22.2 No. 94.1-07 (R2012)
CAN/CSA-C22.2 No. 94.2-07 (R2012)
CAN/CSA-C22.2 No.60079-0:11
CAN/CSA-C22.2 No.60079-15:12
CAN/CSA-C22.2 No.61010-1-12
CAN/CSA C22.2 No. 60529-05 (R2010)
• Ratings
Ex nA nL IIC T5
• Enclosure
Type4X, IP65
• Specific condition of use
Precautions shall be taken to minimize the risk from electrostatic discharge of painted
parts and non-metallic parts of the enclosure.
2) Name Plate
Warning
(Avertissement)
Reference to
the control drawing
number
Ex marking
READ/ LISEZ IM mmmm
Specified model code
Specified suffix code
Specified style code
Specified ambient temperature range
Specified supply voltage and wattage
Serial number
Ex nA nL IIC T5
TYPE 4X, IP65
Warning:
DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE MAY BE
PRESENT
POTENTIAL ELECTROSTATIC CHARGING HAZARD
Avertissement: NE PAS OUVRIR EN ATMOSPHÈRE EXPLOSIVE
DANGER : ELECTRICITÉ STATIQUE
MODEL:
SUFFIX:
STYLE:
AMB.TEMP:
SUPPLY:
NO.:
Ex marking:
Note: “mmmm” means document number of user ’s manual.
Rev.1: December 24, 2015
Doc. No.:
NFM030-A72 P.1
Yokogawa Electric Corporation
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-29
Model: YH8000
Date: January 23, 2015
3) Installation
•
•
•
•
•
•
•
The installation of the equipment should be carried out by the engineers or other professional
personnel of the related expertise. The installation should not be carried out by operators or other
unprofessional personnel.
The installation must be in accordance with Canadian Electrical Code Part I (C22.1) and relevant
local codes.
The equipment must only be used in an area of not more than pollution degree 2, as defined in IEC
60664-1.
When the cable glands and/or adapters are used, they must be Ex “d”, Ex “e”, or Ex “n” type, and
must also meet the ingress protection requirement of IP54 or more.
The thread type of the cable entries shall be confirmed according to the user’s manual.
Screws of the field wiring terminals must be tightened with the specified torque value:
0.22 to 0.25 Nm.
Field wiring for ethernet communication must be in accordance with IEEE 802.3 so as
to avoid overvoltage of > 119 V.
WARNING / AVERTISSEMENT
--------------------------------------------------------------------------------------------------------------------------------------- DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE MAY BE PRESENT
- POTENTIAL ELECTROSTATIC CHARGING HAZARD - READ IM mmmm
- NE PAS OUVRIR EN ATMOSPHÈRE EXPLOSIVE
- DANGER : ELECTRICITÉ STATIQUE - LISEZ IM mmmm
--------------------------------------------------------------------------------------------------------------------------------------------Note: “mmmm” means document number of user ’s manual.
4) Operation
WARNING
--------------------------------------------------------------------------------------------------------------------------------------------• Electrostatic charge may cause an explosion hazard. Avoid any actions that cause the generation of
electrostatic charge, such as rubbing with a dry cloth on coating face of product.
• Take care not to generate mechanical sparking when access to the instrument and peripheral
devices in hazardous locations.
--------------------------------------------------------------------------------------------------------------------------------------------5) Maintenance and Repair
WARNING
--------------------------------------------------------------------------------------------------------------------------------------------• The instrument modification or parts replacement by other than authorized Representative of
Yokogawa Electric Corporation is prohibited and will void the certification.
--------------------------------------------------------------------------------------------------------------------------------------------• When opening the cover, the enclosure should be dry and clean to prevent from ingress water or
dust.
Rev.1: December 24, 2015
Doc. No.:
NFM030-A72 P.2
Yokogawa Electric Corporation
Note: If YH8000 is installed in where more than IP54 is required as ingress protection, the cable gland and/or adapters should be meet
the requirement.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-30
l YH8000-S2 (ATEX Declaration)
(1) Technical Data
•
•
Applicable standards
EN 60079-0: 2012+A11:2013
EN 60079-11: 2012
EN 60079-15: 2010
Ratings
II 3 G Ex nA ic IIC T5 Gc
Enclosure
IP65 (In accordance with EN 60529)
Specific condition of use
Precautions shall be taken to minimize the risk from electrostatic discharge of painted parts
and non-metallic parts of the enclosure.
(2) Name Plate
•
•
(3) Installation
• The installation of the equipment should be carried out by the engineers or other
professional personnel of the related expertise. The installation should not be carried out by
operators or other unprofessional personnel.
• The installation must be in accordance with EN 60079-14 and relevant local codes.
• The equipment must only be used in an area of not more than pollution degree 2, as defined
in IEC 60664-1.
• When the cable glands and/or adapters are used, they must be Ex “d”, Ex “e”, or Ex “n”
type, and must also meet the ingress protection requirement of IP54 or more. If YH8000 is
installed in where more than IP54 is required as ingress protection, the cable gland and/or
adapters should be meet the requirement.
• The thread type of the cable entries shall be confirmed according to the user’s manual.
• Screws of the field wiring terminals must be tightened with the specified torque value: 0.22
to 0.25 Nm.
• Field wiring for ethernet communication must be in accordance with IEEE 802.3 so as to
avoid overvoltage of > 119 V.
WARNING
- DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE MAY BE PRESENT
- POTENTIAL ELECTROSTATIC CHARGING HAZARD
- FOR INSTALLATION AND SAFE USE, READ IM11Y01D01-01
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-31
(4) Operation
WARNING
• Electrostatic charge may cause an explosion hazard. Avoid any actions that cause the
generation of electrostatic charge, such as rubbing with a dry cloth on coating face of
product.
• Take care not to generate mechanical sparking when access to the instrument and
peripheral devices in hazardous locations.
(5) Maintenance and Repair
WARNING
The instrument modification or parts replacement by other than authorized Representative of
Yokogawa Electric Corporation is prohibited and will void the certification.
When opening the cover, the enclosure should be dry and clean to prevent from ingress water or
dust.
l YH8000-E2 (IECEx Certification)
(1) Technical Data
•
•
•
•
Applicable standards
IEC 60079-0: 2011
IEC 60079-11: 2011
IEC 60079-15: 2010
Ratings
Ex nA ic IIC T5 Gc
Enclosure
IP65 (In accordance with IEC 60529)
Specific condition of use
Precautions shall be taken to minimize the risk from electrostatic discharge of painted parts
and non-metallic parts of the enclosure.
(2) Installation
• The installation of the equipment should be carried out by the engineers or other
professional personnel of the related expertise. The installation should not be carried out by
operators or other unprofessional personnel.
• The installation must be in accordance with IEC 60079-14 and relevant local codes.
• The equipment must only be used in an area of not more than pollution degree 2, as defined
in IEC 60664-1.
• When the cable glands and/or adapters are used, they must be Ex “d”, Ex “e”, or Ex “n”
type, and must also meet the ingress protection requirement of IP54 or more. If YH8000 is
installed in where more than IP54 is required as ingress protection, the cable gland and/or
adapters should be meet the requirement.
• The thread type of the cable entries shall be confirmed according to the user’s manual.
• Screws of the field wiring terminals must be tightened with the specified torque value: 0.22
to 0.25 Nm.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
App.8-32
• Field wiring for ethernet communication must be in accordance with IEEE 802.3 so as to
avoid overvoltage of > 119 V.
WARNING
- DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE MAY BE PRESENT
- POTENTIAL ELECTROSTATIC CHARGING HAZARD
- FOR INSTALLATION AND SAFE USE, READ IM11Y01D01-01
(3) Operation
WARNING
• Electrostatic charge may cause an explosion hazard. Avoid any actions that cause the
generation of electrostatic charge, such as rubbing with a dry cloth on coating face of
product.
• Take care not to generate mechanical sparking when access to the instrument and
peripheral devices in hazardous locations.
(4) Maintenance and Repair
WARNING
The instrument modification or parts replacement by other than authorized Representative of
Yokogawa Electric Corporation is prohibited and will void the certification.
When opening the cover, the enclosure should be dry and clean to prevent from ingress water or
dust.
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
TDLS8000
Tunable Diode Laser Spectrometer
Customer
Maintenance
Parts List
9
● Sensor Control Unit
8
11
6
SCU PCB Assy
10
7
8
2
4
3
1
7
5
● Laser Unit
Laser Module
12
13
6
9
8
LU PCB Assy
11
10
7
1
2
14
4
5
Item
1
2
3
4
5
Part No.
K9771CB
K9771JN
A1624EF
B1093BS
K9771KV
Qty
1
1
1
1
1
Description
Cover Assy (with O-ring)
O-ring
Fuse (for SCU)
Set Screw
Hex. L-Key
6
7
8
9
K9771JR
K9771JS
K9772TJ
K9772RA
K9772RB
1
1
1
1
1
Captive Screw
Screw 3pcs Set
O-ring
Process Window Assy (for -X1, -X2)
Process Window Assy (for -H1)
K9772RC
K9772RD
K9772RE
K9772RG
K9772RJ
1
1
1
1
1
Process Window Assy (for -A1, -A2)
Process Window Assy (for -H3)
Process Window Assy (for -C3, -C4)
Process Window Assy (for -C5, -D5)
Process Window Assy (for -L1)
K9772RK
1
Process Window Assy (for -C1, -S1, -D1, -H4)
*:
Item Part No.
10 K9772TH
11 * K9772NA
K9772NB
K9772NC
K9772ND
Qty
1
1
1
1
1
Description
O-ring
Alignment Flange Assy (for -U2)
Alignment Flange Assy (for -U3)
Alignment Flange Assy (for -U4, -LA)
Alignment Flange Assy (for -D5)
12
K9772NE
K9772NF
K9772NG
K9775DA
K9775SB
1
1
1
1
8
Alignment Flange Assy (for -D8)
Alignment Flange Assy (for -J5)
Alignment Flange Assy (for -J8)
FC Almt. Flange Assy (for -FC)
Nut
13
14
K9771NA
A1633EF
1
1
LAO Assy (for -LA)
Fuse (for LU)
Process window assy is not included.
All Rights Reserved. Copyright © 2015 Yokogawa Electric Corporation.
Subject to change without notice.
CMPL 11Y01D01-01EN
1st Edition: Jul. 2015 (YK)
2nd Edition: Jun. 2016 (YK)
Customer
Maintenance
Parts List
K9772XA, K9772XB, K9772XC, K9772XD,
K9772XE, K9772XF, K9772XG, K9772XH,
K9772XJ, K9772XL, K9772XM
Calibration Cell for TDLS8000
3
1
4
2
Item
Part No.
Qty
1
2
3
K9771JS
K9772TJ
K9772RA
K9772RB
K9772RC
1
1
1
1
1
Screw 3pcs Set
O-ring
Process Window Assy (for K9772XA, K9772XB)
Process Window Assy (for K9772XC)
Process Window Assy (for K9772XD)
K9772RD
K9772RE
K9772RG
K9772RJ
K9772RK
1
1
1
1
1
Process Window Assy (for K9772XE)
Process Window Assy (for K9772XF, K9772XG)
Process Window Assy (for K9772XH)
Process Window Assy (for K9772XJ)
Process Window Assy (for K9772XL, K9772XM)
K9772TH
1
O-ring
4
Description
Note: Qty shows necessary number for one side.
All Rights Reserved. Copyright © 2015 Yokogawa Electric Corporation.
Subject to change without notice.
CMPL 11Y01D01-21EN
1st Edition: Jul. 2015 (YK)
2nd Edition: Jun. 2016 (YK)
Customer
Maintenance
Parts List
YH8000
HMI Interface Unit for TDLS8000
1
2
3
6
5
4
Item
Part No.
Qty
1
2
3
4
5
K9774CR
Y9500WU
Y9102XA
A1633EF
A3433JQ
4
4
4
1
1
Screw
Washer
O-ring for screw
Fuse
Power connector
Description
6
A1633JZ
1
RJ45 connector
All Rights Reserved. Copyright © 2015 Yokogawa Electric Corporation.
Subject to change without notice.
CMPL 11Y01D10-01EN
1st Edition: Jul. 2015 (YK)
2nd Edition: Oct. 2015 (YK)
IF8000
Isolation Flange for TDLS8000
Customer
Maintenance
Parts List
5
3
6
(Process connection)
4
2
1
(Analyzer connection)
Item
Part No.
Qty
1
K9775TC
K9775TE
K9775SC
K9775SF
K9775TA
1
1
1
1
1
Nut 4pcs Set (for Analyzer connection “-21”)
Nut 4pcs Set (for Analyzer connection “-50”)
Gasket (for Analyzer connection “-21”)
Gasket (for Analyzer connection “-50”)
Screw 6pcs Set
K9775EA
K9775EB
K9775EC
K9775ED
K9775EE
1
1
1
1
1
Sapphire Window Assy (for Sapphire window type “-12”)
Sapphire Window Assy (for Sapphire window type “-13”)
Sapphire Window Assy (for Sapphire window type “-14”)
Sapphire Window Assy (for Sapphire window type “-15”)
Sapphire Window Assy (for Sapphire window type “-16”)
2
3
4
5
6
Description
K9775EG
K9775EH
K9775EJ
K9775GE
K9775TC
1
1
1
1
1 or 2
Sapphire Window Assy (for Sapphire window type “-17”)
Sapphire Window Assy (for Sapphire window type “-18”)
Sapphire Window Assy (for Sapphire window type “-20”)
O-ring
Nut 4pcs Set (for Process connection “-21”, “-23”, “-31”, “-41”)
K9775TD
K9775TE
2
1 or 2
Nut 4pcs Set (for Process connection “-33”)
Nut 4pcs Set (for -Process connection “50”, “-80”,” -J5”, “-J8”)
Note:
Qty shows necessary number for one side.
All Rights Reserved. Copyright © 2015 Yokogawa Electric Corporation.
Subject to change without notice.
CMPL 11Y01D11-01EN
1st Edition: Jul. 2015 (YK)
2nd Edition: Jun. 2016 (YK)
Customer
Maintenance
Parts List
YC8000
Flow Cell for TDLS8000
5
4
2
1
3
Item
Part No.
Qty
1
2
3
K9775GK
K9775TA
K9775EA
K9775EB
K9775EC
1
1
1
1
1
O-ring
Screw 6pcs Set
Sapphire Window Assy (for -XX)
Sapphire Window Assy (for -HH)
Sapphire Window Assy (for -NH)
K9775ED
K9775EE
K9775EG
K9775EH
K9775EJ
1
1
1
1
1
Sapphire Window Assy (for -H3)
Sapphire Window Assy (for -CC)
Sapphire Window Assy (for -C2)
Sapphire Window Assy (for -HC)
Sapphire Window Assy (for -MC)
K9775GE
K9775TB
1
1
O-ring
Bolt 4pcs Set
4
5
Description
Note: Qty shows necessary number for one side.
All Rights Reserved. Copyright © 2015 Yokogawa Electric Corporation.
Subject to change without notice.
CMPL 11Y01D12-01EN
1st Edition: Jul. 2015 (YK)
2nd Edition: Jun. 2016 (YK)
i
Revision Information
Manual Title : TDLS8000 Tunable Diode Laser Spectrometer
Manual No. : IM 11Y01D01-01EN
Jun. 2016/3rd Edition
TDLS8000, YH8000 software change (Ver. 1.02.01)
Alarm specifications change, “Clear latched alarms” addition, etc. (pages 2-2, 10-1, 10-2, 10-4, 7-9,
App.4-1)
YH8000 software change (Ver. 1.02.01)
Multilingual correspondence (pages 6-11, 8-18, 9-44, 11-5, App.4-3, App.4-4)
HART DD revision change (Rev.2)
“Clear latched alarms” addition, etc. (pages 5-1, 5-2, 7-5, 9-9, 9-34, 9-36, 9-43, 11-4, App.3-2, App.35, App.3-6)
Correction, specifications change, etc. (pages 2-1 to 2-12, 2-17, 2-18, 3-29, 3-34, 9-18, 9-19, Appendix 8,
CMPL 11Y01D01-01EN, CMPL 11Y01D01-21EN, CMPL 11Y01D11-01EN, CMPL 11Y01D12-01EN)
Dec. 2015/2nd Edition
Addition of Type -D2, -C2, -S2, and -E2 of TDLS8000 and YH8000.
Addition of functions due to TDLS8000 software change (Ver. 1.01.05) (Pages 6-12, 6-22, 11-2, 11-3, 11-6 and 11-7)
Correction (CMPL 11Y01D10-01EN)
Jul. 2015/1st Edition
Newly published
Yokogawa Electric Corporation
2-9-32 Nakacho, Musashino-shi, Tokyo 180-8750, JAPAN
Website: http://www.yokogawa.com/
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
Blank Page
Supplement
User’s
Manual
TDLS8000
Tunable Diode Laser Spectrometer
Thank you for selecting our TDLS8000 Tunable Diode Laser Spectrometer.
Please be notified of the following information on the additional function of the product. This is to
be added to User’s Manual “IM 11Y01D01-01EN 3rd Edition” attached to the product. Read this
information carefully before using the product.
Note
n Change of software for TDLS8000 (Ver. 1.02.02)
Backup function has been available so far for an error occurrence at an analog input. But the backup
function is now available as well , even via Modbus network to restore data of process pressure or
temperature.
l Addition: S
ection 6.1.2 Process Pressure (pp.6-1, 6-2)
(applied also to the section 6.1.3 Process Temperature.)
· In order to input via Modbus network, take notice how the Modbus network should correspond to
the backup function as below.
When backup operation starts (the system
starts restoring data)
When backup function stops
Modbus is off
First pressure value is input via Modbus network
NOTE
Maintain Modbus network connected when the backup function is actiavted. If you want to shut down the
Modbus network regularly, set the backup mode “Disable”.
· If you select “Disable” in the backup mode, the last pressure value you enter remains valid
i.e. in the same status as “Hold”.
· After the power of TDLS8000 turns on, the backup function keeps active until the first
pressure value is received. During this period, when you select “Hold”, or “Disable” in the
backup mode, the pressure values will be on hold at the equivalent of the one obtained
within the Analog input range mentioned below (3) with 4mA.
l Addition: Section 11.4 Hold Register on page. 11-5.
NOTE
As a default setting, when Modbus connection is shut down, the backup operation starts to restore data of
temperature and pressure value. If you want to change or disable this backup function, see the description
on the backup mode in the section 6.1.2.
All Rights Reserved. Copyright © 2017, 2nd Edition: Jun. 5, 2017 (YK)
IM 11Y01D01-01EN 1/7
3rd Edition
n Supplementary note on the use of unit connection cables
K9775WD, K9775WE, K9775WF, K9775WG
If the cable length of inter-unit cable is 25m or over between sensor control unit (SCU) and laser unit (LU),
(cable part number : K9775WD, K9775WE, K9775WF, K9775WG), use BELDEN 7957A (outer diameter :
approx. 8.4 mm) with the following instruction.
n Replacement of previous model TDLS200
Inter-unit cable for previous model TDLS200 (BELDEN 1475A), if the length is 25 m or over, is not
compatible. Measure the length of your cable diameter to assure their model and compatibility. The outer
diameter of BELDEN 1475A is approximately 12 mm, and BELDEN 7957A approximately 8.4 mm.
n The power supply voltage
The power supply terminal (VO terminal) on the sensor control unit (SCU) must meet the following
conditions of power voltage in order to provide sufficient power to the laser unit.
If the cable length is between 40 to 50m, the voltage is greater than 22 V.
If the cable length is between 50 to 60m, the voltage is greater than 23 V.
n Using cable gland
If the cable diameter is too small for the cable gland, apply reducers or other adapters with appropriate
diameter so that the cable can fit into the cable gland.
n Cable specification (varies depending on part number as below)
l
p.2-9 Unit Connection Cable
Construction varies depending on part number.
Part Number
Cable Length
Construction
K9775WA
5m
K9775WB
10 m
Double-shielded (Overall shield and
Individual shields) 4-pair cable
K9775WC
20 m
K9775WD
30 m
K9775WE
40 m
K9775WF
50 m
K9775WG
60 m
Overall shield 4-pair cable
l
p.3-12 Figure 3.12 TDLS8000 Wiring
p.3-15 Figure 3.13 Connecting between the Sensor Control Unit (SCU) and Laser Unit (LU)
The connection cables are not pair-individually shielded cables. Connect their overallshielded cables to a shield wire terminal.
IM 11Y01D01-01EN 2/7
3rd Edition
l
p3-14 Types of Wiring and Cabling
Cable
entry
Cable
type
Nominal cross sectional area,
conditions
Shield
Terminal
Withstand
voltage, flame
resistance
1, LU
Interunit
cable
Separately sold cable exclusive to
TDLS8000
K9775WA to K9775WC
(select according to
cable length)
AWG18
4 pairs
Covering outer diameter
approx. 12 mm
Required
(individual
shields for each
pair and overall
shield)
Wire:
Shield: M4 screw
crimp-on terminal
500 V or more
FT-4
Separately sold cable exclusive to
TDLS8000
K9775WD to K9775WG
(select according to
cable length)
AWG24
4 pairs
Covering outer diameter
approx. 8.4 mm
Required
(overall shield)
n Appendix 1-1 Constructing Unit Constructing Unit
The following supplemental information is applied to those parts, Part number : K9775WD, K9775WE,
K9775WF, K9775WG
l Required components and tools(parts number : K9775WD, K9775WE,
K9775WF, K9775WG)
• Wire cutter
• Wire stripper
• Round crimp-on terminals (for M4 screws, nominal cross sectional area of wire 2 mm2), 2
pcs. FV2-S4 by J.S.T. Mfg. Co.,Ltd. or 170782-1 by TE Connectivity, or equivalent
• Crimp tool for round crimp-on terminals
• Heat shrink tubes
Inner diameter 4 mm, length approx. 110 mm, 2 pcs.(for shielded cables)
Inner diameter 10 mm, length approx. 50 mm, 2 pcs.(for external cable covering)
• Heating gun (for shrinking heat shrink tubes)
l Cable Specifications (parts number : K9775WD, K9775WE, K9775WF,
K9775WG)
• Manufacturer and model :
• Number of pairs :
• Shield type :
• Wire diameter :
• Cable outer diameter:
• Flame resistance :
• Operating temperature range :
• Usage environment :
Belden 7957A
4 pairs
Overall shield 4-pair cable
AWG24
Approx. ø8.4 mm
FT4
-40°C to +75°C
Indoor/outdoor
IM 11Y01D01-01EN 3/7
3rd Edition
l How to terminate the cables (parts number : K9775WD, K9775WE,
K9775WF, K9775WG)
Terminate both ends of the cables in the same manner.
(1) Remove about 120 mm of the outer cover from the ends of the cables.
Be careful not to damage the wires, shields, and the like in the cables.
Outer covering
Shield (metallic, braided)
About 120 mm
(2) Bundle the braided shield into one cable, then remove drain wire and the foil
underneath the shield.
Outer covering
Bundle the braided shield wires into one.
(3) Cover the shield with a heat shrink tube and apply heat treatment. Attach a round
crimp-on terminal to the end of the shield cable. Apply heat treatment to the coverstripped part of cables as well.
Outer covering
Attach a round crimp-on terminal
to the shield wire.
Cover with heat shrink tube.
(4) Strip about 7 mm of covering from the ends of each twisted-pair.
Remove about 7 mm of the covering.
Outer covering
The ends of the twisted-pair cables can be screwed onto the terminal block without the covering.
To crimp a pin terminal to the ends of twisted-pairs, use the following recommended product or equivalent.
0.25 mm2 (AWG24)
Connection wire size:
Recommended product: Weidmüeller H 0.25/10
This completes the cable termination procedure for cables K9775WD, K9775WE, K9775WF, K9775WG
(BELDEN 7957A).
IM 11Y01D01-01EN 4/7
3rd Edition
n NEPSI type were added to those products: TDLS8000 and YH8000.
TDLS8000 Tunable Diode Laser Spectrometer
TDLS8000-N2 NEPSI Type of protection “n”, cable entry: Metric thread, piping: Rc
YH8000 HMI Unit
YH8000-N2
NEPSI Type of protection “n”, Metric thread for cable entry
l TDLS8000-N2 (NEPSI Certification)
(1) Technical Data
• Applicable standards
GB 3836.1-2010
GB 3836.8-2014
GB 12476.1-2013
GB 12476.5-2013
IEC 60079-28: 2015
• Certificate No.
GYJ16.1282X
• Ratings
Ex nA nC [op is T6 Ga] IIC T5 Gc
Ex tD A21 IP66 T100°C
• Enclosure
IP66 (In accordance with GB 4208)
• Specific condition of use
Precautions shall be taken to minimize the risk from electrostatic discharge of painted
parts and non-metallic parts of the enclosure.
In case of the enclosure of TDLS8000 with paint layers, if it is mounted conbustible dust
atmosphere Zone 21, it shall be installed in such a way that the risk from electrostatic
discharges and propagating brush discharges caused by rapid flow of dust is avoided.
(2) Installation
• The installation of the equipment should be carried out by the engineers or other
professional personnel of the related expertise. The installation should not be carried out
by operators or other unprofessional personnel.
• Installation, use and maintenance of the product should comply with the manual and
following standards.
GB 3836.13-2013 “Explosive atmospheres Part 13: Equipment repair, overhaul and
reclamation”
GB 3836.15-2000 “Electrical apparatus for explosive gas atmosphere Part 15: Electrical
installations in hazardous area (other than mines)”
GB 3836.16-2006 “Electrical apparatus for explosive gas atmosphere Part 16:
Inspection and maintenance of electrical installation (other than mines)”
GB 50257-2014 “Code for construction and acceptance of electric equipment on fire and
explosion hazard electrical equipment installation engineering”
GB 15577-2007 “Safety regulations for dust explosion prevention and protection”
GB 12476.2-2010 “Electrical apparatus for use in the presence of conbustible dust Part
2: Selection and installation”
• The equipment must only be used in an area of not more than pollution degree 2, as
defined in GB/T 16935.1.
• Mounting orientation of TDLS8000 is within 30 degrees relative to the horizontal
direction.
• When the cable glands and/or adapters are used, they must be Ex “d”, Ex “e”, or Ex
“n” type, and must also meet the ingress protection requirement of IP54 or more. If
TDLS8000 is installed in where more than IP54 is required as ingress protection, the
cable gland and/or adapters should be meet the requirement.
• The thread type of the cable entries shall be confirmed according to the user’s manual.
IM 11Y01D01-01EN 5/7
3rd Edition
• Screws of the field wiring terminals must be tightened with specified torque values as
follows:
– Sensor Control Unit: 0.5 to 0.6 Nm,
– Laser Unit: 0.4 to 0.5 Nm
• Field wiring for ethernet communication must be in accordance with IEEE 802.3 so as to
avoid overvoltage of > 119 V.
WARNING
DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE MAY BE PRESENT
POTENTIAL ELECTROSTATIC CHARGING HAZARD
FOR INSTALLATION AND SAFE USE, READ IM11Y01D01-01
(3) Operation
WARNING
• Electrostatic charge may cause an explosion hazard. Avoid any actions that cause the
generation of electrostatic charge, such as rubbing with a dry cloth on coating face of
product.
• Take care not to generate mechanical sparking when access to the instrument and
peripheral devices in hazardous locations.
(4) Maintenance and Repair
WARNING
The instrument modification or parts replacement by other than authorized Representative of
Yokogawa Electric Corporation is prohibited and will void the certification.
When opening the cover, the enclosure should be dry and clean to prevent from ingress water
or dust.
l YH8000-N2 (NEPSI Certification)
(1) Technical Data
• Applicable standards
GB 3836.1-2010
GB 3836.4-2010
GB 3836.8-2014
• Certificate No.
GYJ16.1281X
• Ratings
Ex nA ic IIC T5 Gc
• Enclosure
IP65 (In accordance with GB 4208)
• Specific condition of use
Precautions shall be taken to minimize the risk from electrostatic discharge of painted
parts and non-metallic parts of the enclosure.
IM 11Y01D01-01EN 6/7
3rd Edition
(2) Installation
• The installation of the equipment should be carried out by the engineers or other
professional personnel of the related expertise. The installation should not be carried out
by operators or other unprofessional personnel.
• Installation, use and maintenance of the product should comply with the manual and
following standards.
GB 3836.13-2013 “Explosive atmospheres Part 13: Equipment repair, overhaul and
reclamation”
GB 3836.15-2000 “Electrical apparatus for explosive gas atmospheres Part 15:
Electrical installations in hazardous area (other than mines)”
GB3836.16-2006 “Electrical apparatus for explosive gas atmospheres Part 16:
Inspection and maintenance of electrical installation (other than mines)”
GB 50257-2014 “Code for construction and acceptance of electric equipment on fire and
explosion hazard electrical equipment installation engineering”
• The equipment must only be used in an area of not more than pollution degree 2, as
defined in GB/T 16935.1.
• When the cable glands and/or adapters are used, they must be Ex “d”, Ex “e”, or Ex “n”
type, and must also meet the ingress protection requirement of IP54 or more. If YH8000
is installed in where more than IP54 is required as ingress protection, the cable gland
and/or adapters should be meet the requirement.
• The thread type of the cable entries shall be confirmed according to the user’s manual.
• Screws of the field wiring terminals must be tightened with the specified torque value:
0.22 to 0.25 Nm.
• Field wiring for ethernet communication must be in accordance with IEEE 802.3 so as to
avoid overvoltage of > 119 V.
WARNING
- DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE MAY BE PRESENT
- POTENTIAL ELECTROSTATIC CHARGING HAZARD
- FOR INSTALLATION AND SAFE USE, READ IM11Y01D01-01
(3) Operation
WARNING
• Electrostatic charge may cause an explosion hazard. Avoid any actions that cause the
generation of electrostatic charge, such as rubbing with a dry cloth on coating face of
product.
• Take care not to generate mechanical sparking when access to the instrument and
peripheral devices in hazardous locations.
(4) Maintenance and Repair
WARNING
The instrument modification or parts replacement by other than authorized Representative of
Yokogawa Electric Corporation is prohibited and will void the certification.
When opening the cover, the enclosure should be dry and clean to prevent from ingress water
or dust.
IM 11Y01D01-01EN 7/7
3rd Edition