Endress + Hauser Operating Instructions Chloromax CCS142D Digital sensor with Memosens technology for determining free chlorine
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
BA00419C/07/EN/14.15
71298136
Products
Solutions
Services
Operating Instructions
Chloromax CCS142D
Digital sensor with Memosens technology for
determining free chlorine
Chloromax CCS142D
Table of contents
Table of contents
1
Document information . . . . . . . . . 4
1.1
1.2
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
9.5
9.6
9.7
Storing the sensor . . . . . . . . . . . . . . . . . . . 25
Regenerating the sensor . . . . . . . . . . . . . 25
Reconditioning the sensor . . . . . . . . . . . . 26
2
Basic safety instructions . . . . . . . 5
10
Accessories . . . . . . . . . . . . . . . . . . . . 28
2.1
2.2
2.3
2.4
2.5
Requirements for personnel . . . . . . . . . . .
Designated use . . . . . . . . . . . . . . . . . . . . . .
Occupational safety . . . . . . . . . . . . . . . . . .
Operational safety . . . . . . . . . . . . . . . . . . .
Product safety . . . . . . . . . . . . . . . . . . . . . . .
5
5
5
6
6
10.1 Connection accessories . . . . . . . . . . . . . . 28
10.2 Installation accessories . . . . . . . . . . . . . . 28
10.3 Accessories for maintenance and
calibration . . . . . . . . . . . . . . . . . . . . . . . . . 28
3
Device description, function . . . 6
3.1
3.2
3.3
Measuring principle . . . . . . . . . . . . . . . . . . 6
Effects on the measuring signal . . . . . . . . 7
Sensor design . . . . . . . . . . . . . . . . . . . . . . . 9
11.1 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
11.2 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4
Incoming acceptance and
product identification . . . . . . . . . 11
4.1
4.2
4.3
4.4
Incoming acceptance . . . . . . . . . . . . . . . .
Product identification . . . . . . . . . . . . . . . .
Scope of delivery . . . . . . . . . . . . . . . . . . . .
Certificates and approvals . . . . . . . . . . . .
5
Installation . . . . . . . . . . . . . . . . . . . . 13
5.1
5.2
5.3
Installation conditions . . . . . . . . . . . . . . . 13
Mounting the sensor . . . . . . . . . . . . . . . . 15
Post-installation check . . . . . . . . . . . . . . . 16
11
11
12
12
6
Electrical connection . . . . . . . . . . 17
6.1
6.2
6.3
Connecting the sensor . . . . . . . . . . . . . . . 17
Ensuring the degree of protection . . . . . 17
Post-connection check . . . . . . . . . . . . . . . 18
7
Commissioning . . . . . . . . . . . . . . . . 19
7.1
7.2
7.3
Function check . . . . . . . . . . . . . . . . . . . . . 19
Sensor polarization . . . . . . . . . . . . . . . . . . 19
Sensor calibration . . . . . . . . . . . . . . . . . . . 19
8
Diagnostics and
troubleshooting . . . . . . . . . . . . . . . 21
9
Maintenance . . . . . . . . . . . . . . . . . . 23
9.1
9.2
9.3
9.4
Maintenance interval . . . . . . . . . . . . . . . .
Clean sensor . . . . . . . . . . . . . . . . . . . . . . .
Replacing the membrane . . . . . . . . . . . . .
Refilling the electrolyte . . . . . . . . . . . . . .
Endress+Hauser
11
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . 29
12
Technical data . . . . . . . . . . . . . . . . . 29
13
Installation and operation in
hazardous environment Class
I Div. 2 . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
23
23
24
24
3
Document information
Chloromax CCS142D
1
Document information
1.1
Warnings
Structure of information
LDANGER
Causes (/consequences)
Consequences of non-compliance (if
applicable)
‣ Corrective action
LWARNING
Causes (/consequences)
Consequences of non-compliance (if
applicable)
‣ Corrective action
LCAUTION
Causes (/consequences)
Consequences of non-compliance (if
applicable)
‣ Corrective action
NOTICE
Meaning
This symbol alerts you to a dangerous situation.
Failure to avoid the dangerous situation will result in a fatal or serious injury.
This symbol alerts you to a dangerous situation.
Failure to avoid the dangerous situation can result in a fatal or serious injury.
This symbol alerts you to a dangerous situation.
Failure to avoid this situation can result in minor or more serious injuries.
This symbol alerts you to situations which may result in damage to property.
Cause/situation
Consequences of non-compliance (if
applicable)
‣ Action/note
1.2
Symbol
Symbols
Meaning
Additional information, tips
Permitted or recommended
Not permitted or not recommended
Reference to device documentation
Reference to page
Reference to graphic
4
Result of a step
Endress+Hauser
Chloromax CCS142D
2
Basic safety instructions
2.1
Requirements for personnel
Basic safety instructions
Installation, commissioning, operation and maintenance of the measuring system may be
carried out only by specially trained technical personnel.
‣ The technical personnel must be authorized by the plant operator to carry out the specified
activities.
‣ The electrical connection may be performed only by an electrical technician.
‣ The technical personnel must have read and understood these Operating Instructions and
must follow the instructions contained therein.
‣ Measuring point faults may be repaired only by authorized and specially trained personnel.
Repairs not described in the Operating Instructions provided must be carried out only
directly at the manufacturer's site or by the service organization.
2.2
Designated use
2.2.1
General
Drinking water, process water and bathing water must be disinfected through the addition of
appropriate disinfectants such as chlorine gas or inorganic chlorine compounds. The dosing
quantity involved must be adapted to continuously fluctuating operating conditions. If the
concentrations in the water are too low, this could jeopardize the effectiveness of the
disinfection process. On the other hand, concentrations which are too high can lead to signs of
corrosion and have an adverse effect on taste, as well as generating unnecessary costs.
The sensor was specifically developed for this application and is designed for continuous
measurement of free chlorine in water. In conjunction with measuring and control equipment,
it allows optimum control of the disinfection process.
2.2.2
Hazardous environment in accordance with cETLus NI Cl. I, Div. 2 1)
‣ Please note the control drawing and the specified application conditions in the appendix of
these Operating Instructions and follow the instructions.
2.2.3
Non-designated use
Use of the device for any purpose other than that described, poses a threat to the safety of
people and of the entire measuring system and is therefore not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.
2.3
Occupational safety
As the user, you are responsible for complying with the following safety conditions:
• Installation guidelines
• Local standards and regulations
1)
Only if connected to CM444R-EA* or CM448R-EA*
Endress+Hauser
5
Device description, function
Chloromax CCS142D
Electromagnetic compatibility
• The product has been tested for electromagnetic compatibility in accordance with the
applicable European standards for industrial applications.
• The electromagnetic compatibility indicated applies only to a product that has been
connected in accordance with these Operating Instructions.
2.4
Operational safety
1.
Before commissioning the entire measuring point, verify that all connections are
correct. Ensure that electrical cables and hose connections are undamaged.
2.
Do not operate damaged products, and safeguard them to ensure that they are not
operated inadvertently. Label the damaged product as defective.
3.
If faults cannot be rectified:
Take the products out of operation and safeguard them to ensure that they are not
operated inadvertently.
2.5
Product safety
The product is designed to meet state-of-the-art safety requirements, has been tested, and
left the factory in a condition in which it is safe to operate. The relevant regulations and
European standards have been observed.
3
Device description, function
3.1
Measuring principle
Free chlorine is determined as hypochlorous acid according to the amperometric measuring
principle.
The hypochlorous acid (HOCl) contained in the medium diffuses through the sensor
membrane and is reduced to chloride ions (Cl-) at the gold cathode. At the silver anode, silver
is oxidized to silver chloride. Electron donation at the gold cathode and electron acceptance at
the silver anode causes a current to flow which is in proportion to the concentration of free
chlorine in the medium at constant conditions.
The concentration of hypochlorous acid depends on the pH value. This dependency can be
compensated for by measuring the pH value in the flow assembly.
The transmitter uses the current signal to calculate the measured variable for concentration in
mg/l.
6
Endress+Hauser
Chloromax CCS142D
Device description, function
3.2
Effects on the measuring signal
3.2.1
pH value
pH-dependency
Molecular chlorine (Cl2) is present at pH values < 4. Consequently, hypochlorous acid (HOCl)
and hypochlorite (OCl– remain within the range of pH 4 to 11 as components of free chlorine.
As hypochlorous acid splits up (dissociates) with an increasing pH value to form hypochlorite
ions (OCl–) and hydrogen ions (H+), the amounts of the individual components of free
effective chlorine change with the pH value. For example, if the amount of hypochlorous acid
is 97% at pH 6, it drops to approx. 3% at pH 9.
100
A
0
%
80
20
60
40
OCl-
HOCl
C
40
60
B
20
80
%
100
0
4
5
6
7
8
pH
9
10
11
A0002017
1
A
B
C
Principle of pH compensation
Measured value with pH compensation
Measured value without pH compensation
pH compensation
For amperometric measurement using the CCS142D chlorine sensor, only the amount of
hypochlorous acid is selectively measured. This works as a powerful disinfectant in a watery
solution. In contrast to this, hypochlorite is an extremely weak disinfectant. Therefore, when
used as a disinfectant at higher pH values, the effectiveness of chlorine is limited. As
hypochlorite ions cannot permeate the sensor membrane, the sensors do not record this
value.
Endress+Hauser
7
Device description, function
Chloromax CCS142D
pH compensation of chlorine sensor signal
To calibrate and verify the chlorine measuring system, a colorimetric reference measurement
must be carried out using the DPD method. Free chlorine reacts with diethyl-pphenylendiamine to form a red dye. The intensity of the red color increases proportionally to
the chlorine content. With the DPD method, the sample is buffered to a pH value of approx
6.3. Therefore, the pH value of the sample is not included in the DPD measurement. Due to
the buffer function in the DPD method, all components of free effective chlorine are recorded
and thus the total free chlorine is measured.
If pH compensation is switched on in the transmitter, the sum of hypochlorous acid (HOCl)
and hypochlorite corresponding to the DPD measurement is calculated from the chlorine
sensor's measuring signal that corresponds to hypochlorous acid (HOCl) and by taking into
account the pH value in the range of pH 4 to 9. For this calculation, the curve is stored in the
transmitter.
When free chlorine is measured with pH compensation switched on, always perform
calibration in pH-compensated mode.
When you use pH compensation, the measured chlorine value that is displayed and applied to
the device output corresponds to the DPD measured value even if the pH values fluctuate. If
you do not use pH compensation, the measured chlorine value corresponds to the DPD
measurement only if the pH value remains unchanged compared with the calibration.
Without pH compensation, the chlorine measuring system must be recalibrated when the pH
value changes.
Accuracy of pH compensation
The accuracy of the pH-compensated measured chlorine value is derived from the sum of
several individual deviations (chlorine, pH, temperature, DPD measurement etc.).
High levels of hypochlorous acid (HOCl) during chlorine calibration have a positive effect on
accuracy, whereas low levels of hypochlorous acid have a negative effect. The inaccuracy of
the pH-compensated measured chlorine value increases the greater the pH difference
between measuring mode and chlorine calibration or the more inaccurate the underlying
individual measured values are.
Calibration taking into account the pH value
The reference measurement (DPD method, photometer) determines the total free chlorine by
buffering to pH 6.3. In contrast to this, amperometric measurement determines only the HOCl
component.
During operation, pH compensation is effective up to a pH value of 9. However, there is hardly
any HOCl left at this pH value, and the measured current is very low. At this point, pH
compensation has the effect of increasing the measured HOCl value to the actual value of the
free chlorine. Calibration of the complete measuring system makes sense only up to a
pH value of the medium of 8 or 8.2.
Sensor
pH value
HOCl content
Uncompensated value
Compensated value
CCS142D-G
8.2
15 %
12 nA
80 nA
CCS142D-A
8.0
20 %
4 nA
20 nA
8
Endress+Hauser
Chloromax CCS142D
Device description, function
Above these pH values, the total error of the measuring system is unacceptably high.
3.2.2
Flow
The minimum flow velocity of the membrane-covered sensor is 15 cm/s (0.5 ft/s). When
using the CCA250 flow assembly, this value corresponds to a flow rate of 30 l/h (8 gal/h)
(upper edge of float at level of red bar mark).
At higher flow rates, the measuring signal is virtually flow-independent. However, if the flow
rate falls below the specified value, the measuring signal depends on the flow. If an INS
proximity switch is installed in the assembly, this unpermitted operating status can be reliably
detected. This triggers an alarm or, if necessary, results in the dosing process being switched
off.
3.2.3
Temperature
Changes in the medium temperature also influence the measuring signal. An increase in
temperature causes the measured value to increase (approx. 4 % per K), while a reduction in
temperature causes the measured value to decrease. Therefore, the temperature should
remain constant after calibration. If the temperature has changed, a recalibration must be
performed.
The Chloromax CCS142D, when used with the Liquiline, enables automatic temperature
compensation (ATC). In this case, the temperature does not need to remain constant, and a
recalibration in the event of temperature changes is not required.
3.3
Sensor design
The sensor consists of the following function units:
• Measuring chamber
– to protect the anode and cathode from the medium
– with large volume of electrolyte for a long service life in combination with the large anode
and the small cathode
• Sensor shaft with
– large anode
– cathode embedded in plastic
– Temperature sensor
• Membrane cap with
– robust PTFE membrane
– special support grid between cathode and membrane for a specified and constant
electrolyte film and thus a relatively constant indication even at varying pressures and
flows
Endress+Hauser
9
Device description, function
Chloromax CCS142D
1
2
1
Memosens plug-in head
2
O-ring
3
Large anode, silver/silver chloride
4
Gold cathode
5
Measuring chamber
6
Membrane cap with dirt-repellent membrane
7
Screw cap for securing the membrane cap
5
3
6
4
7
A0026479
2
10
Sensor structure
Endress+Hauser
Chloromax CCS142D
Incoming acceptance and product identification
4
Incoming acceptance and product identification
4.1
Incoming acceptance
1.
Verify that the packaging is undamaged.
Notify your supplier of any damage to the packaging.
Keep the damaged packaging until the matter has been settled.
2.
Verify that the contents are undamaged.
Notify your supplier of any damage to the delivery contents.
Keep the damaged products until the matter has been settled.
3.
Check the delivery for completeness.
Check it against the delivery papers and your order.
4.
Pack the product for storage and transportation in such a way that it is protected
against impact and moisture.
The original packaging offers the best protection.
The permitted ambient conditions must be observed (see "Technical data").
If you have any questions, please contact your supplier or your local sales center.
4.2
Product identification
4.2.1
Nameplate
The nameplate provides you with the following information on your device:
• Manufacturer identification
• Order code
• Extended order code
• Serial number
• Ambient and process conditions
• Safety information and warnings
• Ex labeling on hazardous area versions
Compare the data on the nameplate with your order.
4.2.2
Product identification
Product page
www.endress.com/ccs142d
Interpreting the order code
The order code and serial number of your product can be found in the following locations:
• On the nameplate
• In the delivery papers
Obtaining information on the product
1.
Go to the product page for your product on the Internet.
Endress+Hauser
11
Incoming acceptance and product identification
Chloromax CCS142D
2.
In the navigation area on the right-hand side, select "Check your device features" under
"Device support".
An additional window opens.
3.
Enter the order code from the nameplate into the search field.
You will receive information on each feature (selected option) of the order code.
4.3
Scope of delivery
The scope of delivery comprises:
• Chlorine sensor
• Bottle containing electrolyte (50 ml)
• Replacement cartridge with pretensioned membrane
• Operating Instructions
4.4
Certificates and approvals
4.4.1
mark
Declaration of Conformity
The product meets the requirements of the harmonized European standards. As such, it
complies with the legal specifications of the EC directives. The manufacturer confirms
successful testing of the product by affixing to it the mark.
4.4.2
Ex approvals 2)
cETLus NI Cl. I, Div. 2
• Conforms with UL STD 61010-1, ANSI/ISA STD 12.12.01, FM STD 3600 and FM STD 3611
• Certified according to CSA STD C22.2 NO. 61010-1 and CSA STD C22.2 NO. 213
• ETL control number: 5000765
• Control drawing: 403936
2)
12
Only if connected to CM444R-EA* or CM448R-EA*
Endress+Hauser
Chloromax CCS142D
Installation
5
Installation
5.1
Installation conditions
5.1.1
Orientation
Do not install overhead!
‣ Install the sensor with an angle of inclination of at least 15 °to the horizontal in an
assembly, support or suitable process connection.
‣ Other angles of inclination are not permitted.
‣ Observe the instructions for installing sensors in the Operating Instructions for the
assembly used.
15°
A
A
Permitted orientation
B
Forbidden orientation
15°
B
A0026480
Endress+Hauser
13
Installation
5.1.2
Chloromax CCS142D
Dimensions
148 (5.83)
91.5 (3.60)
7.5 (0.30)
NPT 3/4"
Ø25 (0.98)
Ø 31 (1.22)
A0007386
3
14
Dimensions in mm (inch)
Endress+Hauser
Chloromax CCS142D
5.2
Mounting the sensor
5.2.1
Measuring system
Installation
A complete measuring system comprises:
• Chloromax CCS142D chlorine sensor
• Assembly, e.g. Flowfit CCA250
• Measuring cable CYK10
• Transmitter, e.g. Liquiline CM44x or CM44xR
• Optional:
– Extension cable CYK11
– When using assembly CCA250: additional sensor(s), e.g. pH sensor CPS31D
7
1
2
6
3
5
4
A0007341
4
1
2
3
4
5
6
7
Example of a measuring system
Liquiline CM44x transmitter
Power cable for transmitter
Chlorine sensor CCS142D
Flowfit assembly CCA250
Inlet to assembly (outlet on rear, not shown in graphic)
pH sensor CPS71D
Measuring cable CYK10
Endress+Hauser
15
Installation
5.2.2
Chloromax CCS142D
Installation in CCA250 flow assembly
The Flowfit CCA250 flow assembly is designed for installing the sensor It allows a pH and
ORP sensor to be installed, in addition to the chlorine or chlorine dioxide sensor. A needle
valve regulates the flow in the range of 30 to 120 l/h (7.9 to 30 gal/h).
Please note the following during installation:
‣ The flow rate must be at least 30 l/h (7.9 gal/h). If the flow drops below this value or stops
‣
‣
completely, this can be detected by an inductive proximity switch and used to trigger an
alarm with locking of the dosage pumps.
If the medium is fed back into an overflow basin, pipe or similar, the resulting
counterpressure on the sensor may not exceed 1 bar (14.5 psi) and must remain constant.
Negative pressure at the sensor e.g. caused by feedback of the medium to the suction side
of a pump, must be avoided.
Additional installation instructions can be found in the Operating Instructions for the
assembly.
5.2.3
Installation in other flow assemblies
When using other flow assemblies, please ensure the following:
‣ The flow velocity against the membrane is always at least 15 cm/s (0.49 ft/s).
‣ The flow direction is upwards so that transported air bubbles are removed and do not
collect in front of the membrane.
‣ The membrane is struck directly by the flow.
5.2.4
Installation in CYA112 immersion assembly
Alternatively, the sensor can be installed in an immersion assembly with threaded connection
NPT ¾", e.g. CYA112.
Please note the following during installation:
‣ Hold the sensor securely in position and screw the assembly onto the sensor so that it is
handtight. This prevents the cable from twisting and rupturing.
‣ To improve the sealing effect, we recommend you wrap a thin PTFE tape around the thread
for assemblies with an NPT ¾" thread.
Additional installation instructions can be found in the Operating Instructions for the
assembly.
5.3
Post-installation check
1.
Check the membrane to ensure it is sealed and undamaged.
Replace if necessary.
2.
Is the sensor installed in an assembly and not suspended from the cable?
Install the sensor only in an assembly or directly via the process connection.
16
Endress+Hauser
Chloromax CCS142D
6
Electrical connection
Electrical connection
LWARNING
Device is live
Incorrect connection may result in injury or death.
‣ The electrical connection may be performed only by an electrical technician.
‣ The electrical technician must have read and understood these Operating Instructions and
must follow the instructions contained therein.
‣ Prior to commencing connection work, ensure that no voltage is present on any cable.
6.1
Connecting the sensor
The electrical connection of the sensor to the transmitter is performed using the measuring
cable CYK10.
GY GND
YE Com B
GN Com A
WH –
BN +
EY/NG
A0024019
5
Measuring cable CYK10
To extend the cable, use measuring cable CYK11. The maximum cable length is 100 m
(328 ft).
6.2
Ensuring the degree of protection
Only the mechanical and electrical connections which are described in these instructions and
which are necessary for the required, designated use, may be carried out on the device
delivered.
‣ Exercise care when carrying out the work.
Otherwise, the individual types of protection (Ingress Protection (IP), electrical safety, EMC
interference immunity) agreed for this product can no longer be guaranteed due, for example,
to covers being left off or cable (ends) which are loose or insufficiently secured.
Endress+Hauser
17
Electrical connection
6.3
Chloromax CCS142D
Post-connection check
Device condition and specifications
Notes
Are the outside of the sensor, assembly, cable undamaged?
Visual inspection
Electrical connection
Notes
Are the installed cables strain-relieved and not twisted?
Is a sufficient length of the cable cores stripped, and is it positioned
in the terminal correctly?
Check the fit (by pulling gently)
Are all the screws terminals properly tightened?
Tighten
Are all cable entries mounted, tightened and leak-tight?
For lateral cable entries, make sure the cables
loop downwards to allow water to drip off
Are all cable entries installed downwards or mounted laterally?
18
Endress+Hauser
Chloromax CCS142D
7
Commissioning
7.1
Function check
Commissioning
Before first commissioning, check if:
• the sensor is correctly installed
• the electrical connection is correct.
If using an assembly with automatic cleaning, check that the cleaning medium (e.g. water or
air) is connected correctly.
LWARNING
Escaping process medium
Risk of injury from high pressure, high temperatures or chemical hazards
‣ Before applying compressed air to an assembly with cleaning facility, make sure the
connections are correctly fitted.
‣ Do not install the assembly in the process if you cannot make the correct connection
reliably.
7.2
Sensor polarization
The voltage applied between cathode and anode by the transmitter polarizes the surface of
the working electrode. Therefore, after switching on the transmitter with the senor
connected, you must wait until the polarization period has elapsed before starting calibration.
To achieve a stable display value, the sensor requires the following polarization periods:
First commissioning
CCS142D-A:
60 min.
CCS142D-G:
90 min.
Recommissioning
CCS142D-A:
30 min.
CCS142D-G:
45 min.
7.3
Sensor calibration
Reference measurement according to the DPD method
To calibrate the measuring system, carry out a colorimetric comparison measurement in
accordance with the DPD method. Chlorine reacts with diethyl-p-phenylendiamine (DPD)
producing a red dye, the intensity of the red color being proportional to the chlorine content.
The intensity of the red color is measured by a photometer (e.g. CCM182) and displayed as
chlorine content.
Endress+Hauser
19
Commissioning
Chloromax CCS142D
Prerequisites
The sensor reading is stable (no drifts or unsteady values for at least 5 minutes). This is
normally guaranteed once the following preconditions have been met:
• The polarization period has elapsed.
• The flow is constant and within the correct range.
• The sensor and the medium are at the same temperature.
• The pH value is within the permitted range.
Zero point adjustment
A zero point adjustment is not required due to the zero point stability of the membranecovered sensor.
However, if you wish to do so, you may perform a zero point adjustment by operating the
sensor in chlorine-free water for at least 15 min..
Slope calibration
1.
Ensure that the pH value and temperature of the medium are constant.
2.
Take a sample for DPD measurement. This must be done in close proximity to the
sensor. Use the sampling tap if available.
3.
Determine the chlorine content using the DPD method.
4.
Enter the measured value into the transmitter (see Operating Instructions for
transmitter).
5.
Using the DPD method, check the calibration several hours or 24 hours later.
You must carry out a slope calibration every time the electrolyte or membrane is
changed!
20
Endress+Hauser
Chloromax CCS142D
8
Diagnostics and troubleshooting
Diagnostics and troubleshooting
When troubleshooting, you must take account of the entire measuring system. This
comprises:
• Transmitter
• Electrical connections and lines
• Assembly
• Sensor
The possible causes of error in the following table refer primarily to the sensor. Before
commencing troubleshooting, ensure that the following operating conditions have been met:
• Constant pH value after calibration, not required for measurement in "pH-compensated"
mode
• Constant temperature after calibration, not required for measurement in "temperaturecompensated" mode
• Medium flow rate of at least 30 l/h (7.9 gal/h)(red bar mark when using the CCA250 flow
assembly)
• No use of organic chlorination agents
If the value measured by the sensor differs significantly from that of the DPD method,
you should first consider all possible malfunctions of the photometric DPD method (see
Operating Instructions for photometer). If necessary, repeat the DPD measurement
several times.
Error
Possible cause
Remedy
No display, no sensor current
No supply voltage at the transmitter
Establish mains connection.
Connection cable between sensor and
transmitter interrupted
Establish cable connection.
Measuring chamber is not filled with
electrolyte
Fill measuring chamber (→ 24)
No input flow of medium
Establish flow, clean filter
Polarization of the sensor not yet
completed
Wait for polarization to be completed
(→ 19).
Membrane defective
Replace membrane cap.
Shunt resistance (e.g. moisture
contact) in the sensor shaft
Open measuring chamber, rub gold cathode
dry. If the transmitter display does not return
to zero, there is a shunt present.
Foreign oxidants interfering with
sensor
Examine medium, check chemicals.
Display value too high
Endress+Hauser
21
Diagnostics and troubleshooting
Chloromax CCS142D
Error
Possible cause
Remedy
Display value too low
Measuring chamber not completely
tightened.
Fully tighten measuring chamber or screw
cap
Membrane soiled
Clean membrane
Air bubble in front of membrane
Release air bubble
Air bubble between cathode and
membrane
Open measuring chamber, top up electrolyte,
tap
Input flow of medium too low
Establish correct input flow (→ 7)
Foreign oxidants interfering with DPD
reference measurement
Examine medium, check chemicals (→ 6).
Use of organic chlorination agents
Use agents according to DIN 19643 (water
may need to be replaced beforehand)
Display fluctuates considerably Hole in membrane
External voltage in medium
22
Replace membrane cap.
Measure voltage between the PMC pin and
the protective ground of the measuring
device (both AC and DC ranges). For values
greater than approx. 0.5 V, find and
eliminate external cause
Endress+Hauser
Chloromax CCS142D
9
Maintenance
Maintenance
Take all the necessary precautions in time to ensure the operational safety and reliability of
the entire measuring system.
NOTICE
Effects on process and process control
‣ When carrying out any work on the system, take into account possible repercussions for
process control or the process itself.
‣ For your own safety, only use genuine accessories. With genuine parts, the function,
accuracy and reliability are also ensured after maintenance work.
9.1
Maintenance interval
1.
Check the measurement at regular intervals; depending on the prevailing conditions, at
least once a month.
2.
Clean the sensor if the membrane is visibly soiled (→ 23).
3.
Replace the electrolyte once per season or every 12 months. Depending on the
chlorine content on site, this period can be reduced or extended.
4.
Calibrate the sensor if desired or when necessary (→ 19).
9.2
Clean sensor
LCAUTION
Diluted hydrochloric acid
Hydrochloric acid causes irritation if it comes into contact with the skin or eyes.
‣ When using diluted hydrochloric acid, wear protective clothing such as gloves and goggles.
‣ Avoid splashes.
NOTICE
Chemicals that reduce surface tension
Chemicals that reduce surface tension can penetrate the sensor membrane and cause
measuring errors due to blockage.
‣ Do not use chemicals that reduce surface tension.
If the membrane is visibly soiled, proceed as follows:
1.
Remove the sensor from the flow assembly.
2.
Clean the membrane mechanically with a gentle water jet. Alternatively, place it for
several minutes in 1 to 5% hydrochloric acid without any further chemical additives.
Endress+Hauser
23
Maintenance
9.3
Chloromax CCS142D
Replacing the membrane
1
2
3
A0026509
1.
Unscrew the measuring chamber (pos. 1).
2.
Unscrew the front screw cap (pos. 3).
3.
Remove the membrane cap (pos. 2) and replace it with a CCY14-WP replacement
cartridge.
4.
Refill the measuring chamber with electrolyte CCY14-F (→ 24).
9.4
Refilling the electrolyte
NOTICE
Damage to membrane and electrodes, air bubbles
Possibility of measuring errors to complete failure of the measuring point
‣ Do not touch the membrane or electrodes. Avoid damaging them.
‣ The electrolyte is chemically neutral and is not hazardous to health. Nonetheless, do not
swallow it and avoid contact with the eyes.
‣ Keep the electrolyte bottle closed after use. Do not transfer the electrolyte to other
containers.
‣ Do not store the electrolyte for longer than 1 year. The electrolyte must not be yellow in
color. Note the use-by date on the label.
‣ Avoid forming air bubbles when pouring the electrolyte into the membrane cap.
24
Endress+Hauser
Chloromax CCS142D
Maintenance
Refilling the electrolyte
1.
Unscrew the measuring chamber from the sensor shaft.
2.
Hold the measuring chamber at an angle and pour in approx.
7 to 8 ml (0.24 to 0.27 fl.oz) of electrolyte, up to the internal thread.
3.
Tap the filled measuring chamber several times on a flat surface so that adherent air
bubbles on the inside can detach and rise.
4.
Insert the sensor shaft vertically from above into the measuring chamber.
5.
Slowly tighten the measuring chamber to the stop. While tightening, excess electrolyte
is forced out at the bottom of the sensor.
6.
If necessary, use a cloth to wipe the measuring chamber and screw cap dry.
9.5
Storing the sensor
During short-term interruptions to measurement:
1.
If the assembly is guaranteed not to empty out,
you may leave the sensor in the flow assembly.
2.
If there is a possibility that the assembly may empty out,
Remove the sensor from the assembly.
3.
To ensure that the membrane of the removed sensor remains moist, moisten the inner
sponge in the protective cap and slide the cap onto the measuring chamber.
During longterm interruptions to measurement, particularly if dehydration is possible:
1.
Remove the sensor from the assembly.
2.
Empty the sensor.
3.
Rinse the measuring chamber and the electrode shaft with cold water and leave to dry.
4.
Screw the sensor down loosely and not to the stop, to ensure that the membrane
remains unstressed.
5.
When recommissioning the sensor, proceed according to the "Commissioning" (→ 19)
section.
9.6
Regenerating the sensor
During measurement, the electrolyte in the sensor is gradually exhausted due to chemical
reactions. The gray-brown silver chloride layer that is applied to the anode at the factory
continues to grow during sensor operation. However, this has no effect on the reaction taking
place at the cathode.
A change in the color of the silver chloride layer indicates an effect on the reaction that is
taking place. Therefore, carry out a visual inspection to ensure that the gray-brown color of
the anode has not changed. If the color of the anode has changed, e.g. if it is spotted, white or
silvery, the sensor must be regenerated.
‣ Send it to the manufacturer for regeneration.
Endress+Hauser
25
Maintenance
9.7
Chloromax CCS142D
Reconditioning the sensor
Longterm operation of the sensor (> 3 months) in a chlorine-free medium, i.e. with very low
sensor currents, may lead to deactivation of the sensor. This deactivation is a continuous
process that results in a lower slope and longer response times. After long-term operation in a
chlorine-free medium, the sensor can be reconditioned.
The following materials are required for reconditioning:
• Deminieralized water
• Polishing sheet (see "Accessories", → 28)
• Beaker
• Approx. 100 ml (3.38 fl.oz) of chlorine bleach lye NaOCl, approx. 13 %, pharmaceutical
quality (available at chemical stores or pharmacies)
1
Sensor
2
Gaseous phase of chlorine bleach lye
3
Chlorine bleach lye
a
Distance between sensor and liquid, 5 to 10 mm (0.2 to 0.4 inch)
1
a
2
3
A0026513
1.
Close the medium inlet and outlet and make sure that no liquid can escape from the
assembly.
2.
Remove the sensor from the assembly.
3.
Unscrew the measuring chamber and put it aside.
4.
Polish the sensor's gold cathode using the polishing sheet: Place a wetted strip of the
sheet in your hand, polish the gold cathode on the strip using circular movements, and
rinse the sensor with demineralized water.
5.
If necessary:
Top up the electrolyte in the measuring chamber and screw the measuring chamber
back onto the sensor shaft.
6.
Fill the beaker to approx. 10 mm (0.4 inch) with chlorine bleach lye and position it
safely.
26
Endress+Hauser
Chloromax CCS142D
Maintenance
7.
The sensor must not touch the liquid.
Place the sensor in the gaseous phase approx. 5 to 10 mm (0.2 to 0.4 inch) above the
chlorine bleach lye.
The sensor current will now increase. The absolute value and the speed of increase
depend on the temperature of the chlorine bleach lye.
8.
When the sensor current has reached a value of several hundred nA:
Leave the sensor in this position for approx. 20 min..
9.
If the value of several hundred nA is not reached:
Cover the beaker to avoid a quick air change.
10. Once the 20 min. have elapsed, re-install the sensor in the assembly.
11. Open the medium inlet and outlet again.
The sensor current will now normalize.
After allowing sufficient settling time (no noticeable drift), calibrate the measuring chain.
Endress+Hauser
27
Accessories
10
Chloromax CCS142D
Accessories
The following are the most important accessories available at the time this
documentation was issued. For accessories not listed here, please contact your service or
sales office.
10.1
Connection accessories
CYK10 Memosens data cable
• For digital sensors with Memosens technology
• Product Configurator on the product page: www.endress.com/cyk10
Technical Information TI00118C
Memosens data cable CYK11
• Extension cable for digital sensors with Memosens protocol
• Product Configurator on the product page: www.endress.com/cyk11
Technical Information TI00118C
10.2
Installation accessories
Flowfit CCA250
• Flow assembly for chlorine and pH/ORP sensors
• Product Configurator on the product page: www.endress.com/cca250
Technical Information TI00062C
10.3
Accessories for maintenance and calibration
Photometer
• Photometer for determining chlorine and pH value
• Order no.: 71257946
Service kit CCS14x
• For chlorine sensors CCS140 / CCS141 / CCS142D
• 2 replacement cartridges, electrolyte 50 ml (1.69 fl.oz), polishing sheets
• Order No. 71076921
28
Endress+Hauser
Chloromax CCS142D
11
Repair
11.1
Return
Repair
The product must be returned if repairs or a factory calibration are required, or if the wrong
product was ordered or delivered. As an ISO-certified company and also due to legal
regulations, Endress+Hauser is obliged to follow certain procedures when handling any
returned products that have been in contact with medium.
To ensure swift, safe and professional device returns, please read the return procedures and
conditions at www.endress.com/support/return-material.
11.2
Disposal
The device contains electronic components and must therefore be disposed of in accordance
with regulations on the disposal of electronic waste.
Observe the local regulations.
12
Technical data
12.1
Input
12.1.1
Measured variables
Free chlorine:
12.1.2
Measuring ranges
CCS142D-A:
CCS142D-G:
12.1.3
hypochlorous acid (HOCl)
0.05 to 20 mg/l Cl2 (at 20 °C (68 °F), pH 7.2)
0.01 to 5 mg/l Cl2 (at 20 °C (68 °F), pH 7.2)
Signal current
CCS142D-A:
CCS142D-G:
Endress+Hauser
Approx. 25 nA per mg/l Cl2 (at 20 °C (68 °F), pH
7.2)
Approx. 80 nA per mg/l Cl2 (at 20 °C (68 °F), pH
7.2)
29
Technical data
Chloromax CCS142D
12.2
Performance characteristics
12.2.1
Reference operating conditions
20 °C (68 °F)
pH 7.2
12.2.2
Response time
T90 < 2 min
in applications involving mainly active chlorination
12.2.3
Measured value resolution
CCS142D-A:
CCS142D-G:
12.2.4
Approx. 15 μg/l Cl2
Approx. 5 μg/l Cl2
Measured error
1% of reading
12.2.5
Repeatability
• Sensor: ± 1%
• Reference method: depending on version
Calibration standards do not have long-term stability.
12.2.6
Nominal slope
CCS142D-A:
CCS142D-G:
12.2.7
-25 nA per mg/l
-80 nA per mg/l
Long-term drift
< 1.5 % per month
12.2.8
Polarization time
CCS142D-A:
CCS142D-G:
12.2.9
Commissioning: 60 min, Reoperation: 30 min
Commissioning: 90 min, Reoperation: 45 min
Operating time of the electrolyte
At average medium concentrations of 1 mg/l Cl2
CCS142D-A:
CCS142D-G:
30
> 5 years
> 3 years
Endress+Hauser
Chloromax CCS142D
Technical data
12.2.10 Chlorine intrinsic consumption
At average medium concentrations of 1 mg/l Cl2 and under reference operating conditions
CCS142D-A:
CCS142D-G:
12.3
Environment
12.3.1
Ambient temperature range
25 ng Cl2 per hour
100 ng Cl2 per hour
–5 to 55 ˚C (20 to 130 ˚F), non-freezing
12.3.2
Storage temperature
With electrolyte:
Without electrolyte:
12.3.3
5 to 50 °C (40 to 120 °F)
-20 to 60 °C (-4 to 140 °F)
Degree of protection
IP 68 (up to mounting collar Ø 36 mm (1.42"))
12.4
Process
12.4.1
Process temperature
0 to 45 °C (32 to 110 °F), non-freezing
12.4.2
Process pressure
max. 2 bar (29 psi) absolute, if installed in assembly CCA250
12.4.3
pH range
At average medium concentrations of 1 mg/l Cl2 and under reference operating conditions
Calibration
CCS142D-A:
CCS142D-G:
Measurement
pH 4 to 8
pH 4 to 8.2
pH 4 to 9
Chlorine measurement possible up to pH 9 with limited accuracy
12.4.4
Flow
min. 30 l/h (8 gal/h), in assembly CCA250
12.4.5
Minimum flow
min. 15 cm/s (0.5 ft/s)
Endress+Hauser
31
Installation and operation in hazardous environment Class I Div. 2
12.5
Mechanical construction
12.5.1
Dimensions
Chloromax CCS142D
→ 14
12.5.2
Weight
0.1 kg (0.2 lbs)
12.5.3
Materials
Sensor shaft:
Membrane:
Membrane cap:
Cathode:
Anode:
12.5.4
PVC
PTFE
PBT (GF 30), PVDF
Gold
Silver/silver chloride
Cable specification
max. 100 m (330 ft), incl. Cable extension
13
Installation and operation in hazardous environment
Class I Div. 2
Non-sparking device for use in specified hazardous environment in accordance with:
• cETLus Class I Div. 2
• Gas group A, B, C, D
• Temperature class T6, -5 °C (23 °F) < Ta < 55 °C (131 °F)
• Control drawing: 403936
32
Endress+Hauser
Chloromax CCS142D
Endress+Hauser
Installation and operation in hazardous environment Class I Div. 2
33
Index
Chloromax CCS142D
Index
A
Accessories . . . . . . . . . . . . . . . . . . . . . 28
Ambient temperature range . . . . . . . . . . . 31
C
Cable specification . . . . . . . . . . . . . . . . .
Check
Connection . . . . . . . . . . . . . . . . . .
Function . . . . . . . . . . . . . . . . . . . .
Installation . . . . . . . . . . . . . . . . . .
Chlorine intrinsic consumption . . . . . . . . . .
Cleaning . . . . . . . . . . . . . . . . . . . . . . .
Connection
Check . . . . . . . . . . . . . . . . . . . . .
Ensuring the degree of protection . . . . .
32
18
19
16
31
23
18
17
D
Declaration of Conformity . . . . . . . . . . . . . 12
Degree of protection
Ensuring . . . . . . . . . . . . . . . . . . . . 17
Technical data . . . . . . . . . . . . . . . . 31
Designated use . . . . . . . . . . . . . . . . . . . . 5
Device description . . . . . . . . . . . . . . . . . . 6
Diagnostics . . . . . . . . . . . . . . . . . . . . . 21
Disposal . . . . . . . . . . . . . . . . . . . . . . . 29
E
Effect on the measuring signal
Flow . . . . . . . . . . . . . . . . . . . . . . . 9
pH value . . . . . . . . . . . . . . . . . . . . . 7
Temperature . . . . . . . . . . . . . . . . . . 9
Electrical connection . . . . . . . . . . . . . . . . 17
Electrolyte . . . . . . . . . . . . . . . . . . . . . . 24
Environment . . . . . . . . . . . . . . . . . . . . 31
Ex approvals . . . . . . . . . . . . . . . . . . . . 12
F
Flow . . . . . . . . . . . . . . . . . . . . . . . . 9, 31
Flow assembly . . . . . . . . . . . . . . . . . . . 16
Function . . . . . . . . . . . . . . . . . . . . . . . . 6
Function check . . . . . . . . . . . . . . . . . . . 19
I
Immersion assembly . . . . . . . . . . . . . . . . 16
Incoming acceptance . . . . . . . . . . . . . . . 11
Installation
Check . . . . . . . . . . . . . . . . . . . . . 16
34
Flow assembly . . . . . . . . . . . . . . . .
Immersion assembly . . . . . . . . . . . . .
Orientation . . . . . . . . . . . . . . . . . .
Sensor . . . . . . . . . . . . . . . . . . . . .
Installation instructions . . . . . . . . . . . . . .
16
16
13
15
13
L
Long-term drift . . . . . . . . . . . . . . . . . . . 30
M
Maintenance interval . . . . . . . . . . . . . . . 23
Materials . . . . . . . . . . . . . . . . . . . . . . 32
Measured error . . . . . . . . . . . . . . . . . . . 30
Measured value resolution . . . . . . . . . . . . 30
Measured variables . . . . . . . . . . . . . . . . 29
Measuring principle . . . . . . . . . . . . . . . . . 6
Measuring ranges . . . . . . . . . . . . . . . . . 29
Measuring signal . . . . . . . . . . . . . . . . . . 7
Measuring system . . . . . . . . . . . . . . . . . 15
Minimum flow . . . . . . . . . . . . . . . . . . . 31
N
Nameplate . . . . . . . . . . . . . . . . . . . . . 11
Nominal slope . . . . . . . . . . . . . . . . . . . 30
O
Operating time of the electrolyte . . . . . . . . . 30
Orientation . . . . . . . . . . . . . . . . . . . . . 13
P
Performance characteristics . . . . . . . . . . . 30
pH range . . . . . . . . . . . . . . . . . . . . . . . 31
pH value . . . . . . . . . . . . . . . . . . . . . . . . 7
Polarization time . . . . . . . . . . . . . . . . . . 30
Process . . . . . . . . . . . . . . . . . . . . . . . . 31
Process pressure . . . . . . . . . . . . . . . . . . 31
Process temperature . . . . . . . . . . . . . . . . 31
Product identification . . . . . . . . . . . . . . . 11
R
Reconditioning . . . . . . . . . . . . . . . . . . .
Reference operating conditions . . . . . . . . .
Regeneration . . . . . . . . . . . . . . . . . . . .
Repair . . . . . . . . . . . . . . . . . . . . . . . .
Repeatability . . . . . . . . . . . . . . . . . . . .
Replacing the membrane . . . . . . . . . . . . .
Response time . . . . . . . . . . . . . . . . . . .
26
30
25
29
30
24
30
Endress+Hauser
Chloromax CCS142D
Index
Return . . . . . . . . . . . . . . . . . . . . . . . . 29
S
Safety instructions . . . . . . . . . . . . . . . . . . 5
Scope of delivery . . . . . . . . . . . . . . . . . . 12
Sensor
Calibration . . . . . . . . . . . . . . . . . . 19
Cleaning . . . . . . . . . . . . . . . . . . . . 23
Connecting . . . . . . . . . . . . . . . . . . 17
Design . . . . . . . . . . . . . . . . . . . . . . 9
Mounting . . . . . . . . . . . . . . . . . . . 15
Polarization . . . . . . . . . . . . . . . . . . 19
Reconditioning . . . . . . . . . . . . . . . . 26
Refilling the electrolyte . . . . . . . . . . . 24
Regenerating . . . . . . . . . . . . . . . . . 25
Replacing the membrane . . . . . . . . . . 24
storage . . . . . . . . . . . . . . . . . . . . . 25
Sensor design . . . . . . . . . . . . . . . . . . . . . 9
Storage . . . . . . . . . . . . . . . . . . . . . . . 25
Storage temperature . . . . . . . . . . . . . . . . 31
Symbols . . . . . . . . . . . . . . . . . . . . . . . . 4
T
Technical data
Environment . . . . . . . . . . . . . . . . . 31
Input . . . . . . . . . . . . . . . . . . . . . . 29
Mechanical construction . . . . . . . . . . 32
Performance characteristics . . . . . . . . 30
Process . . . . . . . . . . . . . . . . . . . . . 31
Temperature . . . . . . . . . . . . . . . . . . . . . 9
Troubleshooting . . . . . . . . . . . . . . . . . . 21
U
Use . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
W
Warnings . . . . . . . . . . . . . . . . . . . . . . . 4
Weight . . . . . . . . . . . . . . . . . . . . . . . . 32
Endress+Hauser
35
*71298136*
71298136
www.addresses.endress.com