Endress + Hauser Operating Instructions Oxymax W COS41 Dissolved oxygen sensor
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Operating Instructions
Oxymax W COS41
Dissolved oxygen sensor
BA284C/07/en/04.04
51506691
Brief overview
Here is how to use these Operating Instructions to commission your sensor quickly and safely:
Safety instructions
Page 4 ff.
Page 5
General safety instructions
Explanation of the warning symbols
You can find special instructions at the appropriate position in the chapter in question. The
positions are indicated with the icons Warning #, Caution " and Note !.
Æ
Installation
Page 7 ff.
Here you can find installation conditions such as sensor dimensions and the angle of installation.
Installation examples can be found here.
Page 10 ff.
Æ
Wiring
Page 13 ff.
Refer to theses pages for sensor wiring.
Æ
Sensor design and measuring principle
Page 15
Page 16 ff.
Page 16 ff.
Here you can read about the sensor design.
The measuring principle is explained on this page.
Here you can find the possible calibration methods.
Æ
Commissioning
Page 19 ff.
The sensor must be polarised before calibration. Please, read on the given page how to do it.
Æ
Maintenance
Page 20 ff.
Page 21 ff.
Page 26
It is absolutely essential to carry out maintenance tasks on a regular basis.
Individual parts are subject to normal wear and tear. Here you can find out how to replace such
parts.
Here you can find an overview of the spare parts which can be delivered as well as an overview
of the system.
Æ
Trouble-shooting
Page 24 ff.
If faults occur during operation, use the checklist to locate the cause.
Æ
Æ
Index
Page 29 ff.
You can find important terms and keywords on the individual sections here. Use the keyword
index to find the information you need quickly and efficiently.
Endress+Hauser
Oxymax W COS41
Table of contents
1
Safety instructions . . . . . . . . . . . . . . . . 4
10
Technical data . . . . . . . . . . . . . . . . . . . 27
1.1
1.2
1.3
1.4
1.5
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Installation, commissioning and operation . . . . . . . .
Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . .
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Notes on safety icons and symbols . . . . . . . . . . . . . .
10.1
10.2
10.3
10.4
10.5
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance characteristics . . . . . . . . . . . . . . . . . .
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Mechanical construction . . . . . . . . . . . . . . . . . . . .
2
Identification . . . . . . . . . . . . . . . . . . . . 6
2.1
2.2
Product structure . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3
Installation . . . . . . . . . . . . . . . . . . . . . . 7
3.1
3.2
3.3
3.4
3.5
Incoming acceptance, transport, storage . . . . . . . . . . 7
Installation conditions . . . . . . . . . . . . . . . . . . . . . . . 7
Installation instructions . . . . . . . . . . . . . . . . . . . . . . 8
Installation examples . . . . . . . . . . . . . . . . . . . . . . . 10
Post-installation check . . . . . . . . . . . . . . . . . . . . . . 12
4
Wiring . . . . . . . . . . . . . . . . . . . . . . . . 13
4.1
4.2
4.3
Direct connection to the transmitter . . . . . . . . . . . 13
Connection via junction box . . . . . . . . . . . . . . . . . 13
Post-connection check . . . . . . . . . . . . . . . . . . . . . . 14
5
Operation . . . . . . . . . . . . . . . . . . . . . . 15
5.1
5.2
5.3
Sensor design . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . 16
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
6
Commissioning. . . . . . . . . . . . . . . . . . 19
6.1
6.2
6.3
Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Polarisation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Calibrate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
7
Maintenance. . . . . . . . . . . . . . . . . . . . 20
7.1
7.2
Cleaning the sensor . . . . . . . . . . . . . . . . . . . . . . . . 20
Regenerating the sensor . . . . . . . . . . . . . . . . . . . . . 21
8
Accessories. . . . . . . . . . . . . . . . . . . . . 23
8.1
8.2
8.3
Connection accessories . . . . . . . . . . . . . . . . . . . . . 23
Installation accessories . . . . . . . . . . . . . . . . . . . . . . 23
Measurement, controlling and sensor cleaning . . . . 23
9
Trouble-shooting . . . . . . . . . . . . . . . . 24
9.1
9.2
9.3
9.4
9.5
Trouble-shooting instructions . . . . . . . . . . . . . . . .
Sensor checks . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Endress+Hauser
4
4
4
4
5
27
27
27
27
28
Index . . . . . . . . . . . . . . . . . . . . . . . . . . 29
24
25
26
26
26
3
Safety instructions
Oxymax W COS41
1
Safety instructions
1.1
Designated use
The oxygen sensor COS 41 is suitable for continuous measurement of dissolved oxygen in water.
Typical applications are:
• Measuring oxygen content in activated sludge basins.
The measuring signal is used for monitoring and as a control parameter.
• Checking oxygen content in the outlet of a sewage treatment plant.
• Measuring and controlling the oxygen content in fish farming water.
• Oxygen enrichment in drinking water.
Any other use than the one described here compromises the safety of persons and the entire
measuring system and is, therefore, not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.
1.2
Installation, commissioning and operation
Please note the following items:
• Installation, electrical connection, commissioning, operation and maintenance of the measuring
system must only be carried out by trained technical personnel.
The technical personnel must be authorised for the specified activities by the system operator.
• Technical personnel must have read and understood these Operating Instructions and must
adhere to them.
• Before commissioning the entire measuring point, check all the connections for
correctness. Ensure that electrical cables and hose connections are not damaged.
• Do not operate damaged products and secure them against unintentional
commissioning. Mark the damaged product as being defective.
• Measuring point faults may only be rectified by authorised and specially trained
personnel.
• If faults can not be rectified, the products must be taken out of service and secured against
unintentional commissioning.
• Repairs not described in these Operating Instructions may only be carried out at the
manufacturer’s or by the service organisation.
1.3
Operational safety
The sensor has been designed and tested according to the state of the art and left the factory in
perfect functioning order.
Relevant regulations and European standards have been met.
As the user, you are responsible for complying with the following safety conditions:
• Installation instructions
• Local prevailing standards and regulations.
1.4
Return
If the device requires repair, please send it cleaned to the sales centre responsible. Please use the
original packaging, if possible.
Please enclose the completed "Declaration of contamination" (copy the second last page of these
Operating Instructions) with the packaging and the transportation documents.
No repair without completed "Declaration of contamination"!
4
Endress+Hauser
Oxymax W COS41
Safety instructions
1.5
#
"
!
Endress+Hauser
Notes on safety icons and symbols
Warning!
This symbol alerts you to hazards. They can cause serious damage to the instrument or to persons
if ignored.
Caution!
This symbol alerts you to possible faults which could arise from incorrect operation. They could
cause damage to the instrument if ignored.
Note!
This symbol indicates important items of information.
5
Identification
Oxymax W COS41
2
Identification
2.1
Product structure
Cable length
2 Cable length: 7 m (22.97 ft)
4 Cable length: 15 m (49.22 ft)
8 Without Cable (TOP 68 version only)
9 Special design to customer specifications
Cable connection
F Fixed cable connection
S Cable connection using TOP 68 plug
COS 41-
2.2
Complete order code
Scope of delivery
The following items are included in the delivery:
• Oxygen sensor with transport protection cap for membrane protection
• Accessories Set with the following contents:
– 2 replacement cartridges (replacement membrane caps)
– 10 plastic ampoules containing electrolyte
– 1 sealing kit with 3 O-rings
– 6 abrasive sheets
• Operating Instructions english
If you have any questions, please contact your supplier or your sales centre responsible .
6
Endress+Hauser
Oxymax W COS41
Installation
3
Installation
3.1
Incoming acceptance, transport, storage
Ø40 / 1.57
144 / 5.67
mm / inch
Ø40 / 1.57
C07-COS41xxx-06-05-00-en-001.eps
Fig. 1:
Endress+Hauser
Fixed cable version
190 / 7.48
Dimensions
190 / 7.48
3.2.1
171 / 6.73
Installation conditions
144 / 5.67
3.2
171 / 6.73
• Make sure the packaging is undamaged!
Inform the supplier about damage to the packaging.
Keep the damaged packaging until the matter has been settled.
• Make sure the contents are undamaged!
Inform the supplier about damage to the delivery contents.
Keep the damaged products until the matter has been settled.
• Check that the scope of delivery is complete and agrees with your order and the
shipping documents.
• The packaging material used to store or to transport the product must provide shock protection
and humidity protection. The original packaging offers the best protection. Also, keep to the
approved ambient conditions (see "Technical data").
• If you have any questions, please contact your supplier or your sales centre responsible.
mm / inch
C07-COS41xxx-06-05-00-en-002.eps
Fig. 2:
TOP 68 plug-in head version
7
Installation
Oxymax W COS41
3.2.2
Angle of installation
The sensor can be installed up to the horizontal in an assembly, support or a suitable process
connection (Fig. 3). Other angles are not permissible. Do not install the sensor overhead.
A
C07-COS41xxx-11-05-00-xx-001.eps
Fig. 3:
A
!
Angle of installation
Permissible installation positions: 0 ... 180 °, overhead installation is not permitted
Note!
Make sure you comply with the instructions for installing sensors. You will find them in the
Operating Instructions for the assembly used.
3.2.3
Installation place
• Select the installation location so that there is easy access for later calibration.
• Make sure that upright posts and assemblies are secured safely and vibration-free.
• For immersed operation in an activated sludge basin, select an installation location which
produces a typical oxygen concentration.
3.3
Installation instructions
3.3.1
Measuring system
A complete measuring system comprises at least:
• Oxygen sensor
• Transmitter, e.g. Liquisys M COM 223/253
• Special measuring cable
• Immmersion, flow or retractable assembly
Optional:
• Universal suspension assembly support CYH 101 for immersed operation
• Junction box VBM (with cable extension)
• Automatic spray cleaning system Chemoclean
8
Endress+Hauser
Oxymax W COS41
Installation
1
4
5
2
3
C07-COS41xxx-14-05-00-xx-001.eps
Fig. 4:
1
2
3
4
5
Measuring system (example)
Junction box VBM (optional)
Transmitter Liquisys M COM 253
Special measuring cable
Immersion assembly
Oxygen sensor
3.3.2
!
Installing a measuring point
Note!
For immersed operation, install the individual modules away from the basin on a solid base. Only
carry out the final installation at the intended installation location.
For a complete installation of a measuring point, proceed as follows:
"
#
Endress+Hauser
1.
Install a retractable or a flow assembly (if used) into the process.
2.
Connect the water supply to the rinse connections (if you use an assembly with cleaning
function).
3.
Install and connect the oxygen sensor.
4.
Install an immersion or an suspension assembly (if used) into the process.
Caution!
• For immersed operation, the sensor must be installed in an immersion assembly (e.g. CYA 611).
Do not install the sensor suspended from the cable.
• Screw the sensor into the assembly so that the cable is not twisted.
• Avoid exerting excessive tensile force on the cable (e.g. from jerky pulling).
• Select the installation location so that there is easy access for later calibration.
Warning!
When using metallic assemblies and installation equipment, comply with national grounding
regulations.
9
Installation
Oxymax W COS41
3.4
Installation examples
3.4.1
Immersion operation
Upright post and chain assembly
For large basins, where sufficient installation distance is required from the basin edge (aeration
basin, especially), it is advisable to use the upright post and chain assembly (Fig. 5, Fig. 6). The free
swinging of the immersed assembly practically rules out vibrations from the upright post.
A good self-cleaning of the membrane surface is reached due to the swinging of the assembly.
According to this effect, the sensor life time can be extended.
56/2.20
1
1
3
6/0.24
5
4
60/2.36
120/4.72
22/0.87
6
25/0.98
58/2.58
Ø 40/1.57
Ø 6,5/0.26
450/17.72
950/37.40
4
Ø 14/0.55
1480/58.27
2
Ø 40/1.57
1500/59.06
max. 1400/55.12
1630/64.17
3
1580/64.20
2
Ø 35/1.38 x 1.5/0.06
60/2.36 x 60/2.36 x 2/0.08
4
12
xØ
2
.4
/0
150/5.91
mm/inch
Ø 48/1.89
mm/inch
5
C07-COS41xxx-11-05-00-en-002.eps
Fig. 5:
1
2
3
4
5
6
10
Universal suspension assembly holder with
immersible pendulum assembly
Weather protection cover
Upright post, square pipe SS AISI 304
Transverse pipe SS AISI 304
Star handle
Second fixing possibility for transverse pipe
Immersion assembly CYA 611
C07-COS41xxx-11-05-00-en-004.eps
Fig. 6:
1
2
3
4
5
Immersion assembly CYA 611
Protection cap
Worm drive hose clip
Pipe clips (detailed drawing in right half)
PVC pipe
Threaded coupling
Endress+Hauser
Oxymax W COS41
Installation
Upright post and fixed immersion assembly
The preferable type of installation for strong or turbulent flow (> 0.5 m/s) in the basin or open
channels is to secure the device to an upright post and a securely mounted immersion tube (Fig. 7).
If the flow is very strong, a second transverse pipe can be installed with its own pipe support.
2
1
3
1500/59.06
1
max. 1400/55.12
1490/58.66
120/4.72
4
4 x Ø 12/0.42
2000/78.74 (3500/137.80)
Ø 40/1.57
450/17.72
950/37.40
4
30/1.18 x 30/1.18 x 2/0.06
150/5.91
mm/inch
C07-COS41xxx-11-05-00-en-003.eps
Fig. 7:
1
2
3
4
Universal suspension assembly holder with immersion assembly
Star handle
Pipe holder
Fixing bracket
Immersion assembly (= immersion tube)
Basin rim mounting
For fixing to the sides of the basin or channel, we recommend the use of an basin rim mounting of
the immersion assembly (Fig. 8, Fig. 9).
5
1
5
2
2
4
4
3
3
C07-COS41xxx-11-05-00-xx-005.eps
Fig. 8:
1
2
3
1
Horizontal basin rim mounting
Protection cover for cable entry
Assembly holder
Immersion assembly SS 1.4301 (AISI 304)
C07-COS41xxx-11-05-00-xx-006.eps
Fig. 9:
4
5
Vertical basin rim mounting
Basin rim mounting
Star handle
If there is strong turbulence or flow, a second basin rim mounting must be used for the immersion
assembly.
Endress+Hauser
11
Installation
Oxymax W COS41
Floating body
To aid installation in strongly fluctuating water levels, e.g. in rivers or lakes, there is a floating body
COA 110-50 available (Fig. 10).
1
1
Cable route with strain relief and rain protection
Mounting ring for ropes and chains with locking
screw
Lugs Ø 15; 3x120° for anchoring
Saltwater-resistant plastic float
Pipe 40x1, rust-proof steel SS 1.4571 (AISI 316Ti)
Shock absorber and weight
Oxygen sensor
2
2
3
Ø200
220
3
4
5
4
500
800
6
7
G1
5
6
7
C07-COS41xxx-11-05-00-de-002.eps
Fig. 10:
Floating body
3.4.2
Flow operation
The COA 250-A flow assembly (Fig. 11) with automatic self-venting is suitable for use in pipelines
or hose connections. The inlet is at the bottom of the assembly, the outlet at the top (connection
thread G ¾). It can be installed in a pipe by using two 90° pipe brackets to allow inflow to the
assembly (Fig. 12, Pos. 6).
mm/inch
1
1
G¾
3
196/7.72
Ø34/1.34
2
8
3
2
68/2.68
G¾
4
2.5/0.10
Ø34/1.34
23/0.91
7
90/3.54
100/3.94
5
1
2
3
4
5
3.5
Flow assembly COA 250-A
Screw-in part for sensor
Screw ring
Meter body
Connection thread G¾
Dummy plug (connection for spray head COR 3)
5
4
C07-COS41xxx-11-05-00-xx-009.eps
Fig. 12:
C07-COS41xxx-11-05-00-de-008.eps
Fig. 11:
6
1
2
3
4, 7
5
6
8
Bypass installation with manually
actuated valves or solenoid valves
Main line
Medium return
Oxygen sensor
Manually actuated or solenoid valves
Flow assembly COA 250-A
90 ° pipe bracket
Medium removal
Post-installation check
• Check the membrane for leak tighness und replace it if necessary.
• Compliance with permissible sensor installation position?
• Is the sensor installed to an assembly and is not suspended from the cable?
• Avoid moisture by rain by fitting the protective cap to the immersion assembly?
12
Endress+Hauser
Oxymax W COS41
Wiring
4
#
Wiring
Warning!
• The electrical connection must only be carried out by authorised technical personnel.
• Technical personnel must have read and understood the instructions in this manual and must
adhere to them.
• Ensure that there is no voltage at the power cable before beginning the connection work.
4.1
Direct connection to the transmitter
The sensor COS 41 is connected using a special measuring cable (Fig. 13). The wiring diagram is
contained in the Operating Instructions of the COM 223/253-DX/DS transmitter.
S
RD
WH
GN
12
K
90
A
91
BN
Terminal
Assignment
S
12
90
91
11
Outer screen
Active inner screen (NTC)
Cathode
Anode
NTC temperature sensor
11
C07-COS41xxx-04-05-00-xx-001.eps
Fig. 13:
!
Special measuring cable CYK 71
Note!
The interior white and yellow pilot wires have no function.
4.2
Connection via junction box
To lengthen the sensor connection beyond the length of the fixed cable, you require a junction box
VBM (Fig. 15, Fig. 16). The connection is lengthened to the transmitter using the special measuring
cable CYK 71.
4
2
1
S
S
S
RD
RD
RD
91
90
3
WH
WH
WH
GN
GN
GN
BN
BN
BN
12
11
S
C07-COS41xxx-04-05-00-xx-002.eps
Fig. 14:
1
2
3
4
Endress+Hauser
Connection via junction box VBM
Sensor
Junction box
Extension cable
Transmitter
13
Wiring
Oxymax W COS41
113
125
62,5
Pg 13,5
10
54
136
52
80
C07-COS41xxx-04-05-00-de-003.eps
Fig. 15:
4.3
Junction box VBM, side view
C07-COS41xxx-04-05-00-de-004.eps
Fig. 16:
Junction box VBM, top view
Post-connection check
Instrument status and specifications
Remarks
Are the sensor, assembly, junction box or cable damaged?
Visual inspection
Electrical connection
Remarks
Does the supply voltage of the transmitter match the specifications on the
nameplate?
110/230 V AC
24 V AC/DC
Are the installed cables strain-relieved and not twisted ?
Is the cable type route completely isolated ?
Power cable/weak current cable
Are the power supply and signal cable correctly connected to the
transmitter ?
Use the connection diagram of
COM 2x3.
Are all the screws terminals properly tightened ?
Are all the cable entries installed, tightened and sealed ?
Are all the cable entries installed downwards or lateral ?
14
For cable entries lateral: cable loops
downwards for water to be able to drip off.
Endress+Hauser
Oxymax W COS41
Operation
5
Operation
5.1
Sensor design
10
1
11
2
9
12
8
13
7
3
C07-COS41xxx-16-05-07-xx-001.eps
Fig. 18:
4
5
7
8
9
10
11
12
13
Sensor head, cutaway drawing
Membrane
Electrolyte
Anode
Threaded connection for protection basket
Sealing ring
Threaded connection for membrane cap
Membrane cap
9
8
7
6
6
C07-COS41xxx-03-05-00-xx-001.eps
Fig. 17:
1
2
3
4
5
6
7
8
9
Sensor design
Sensor cable
Threaded connection NPT ¾"
Threaded connection G1
Sensor shaft
Protection basket
Gold cathode
Membrane
Electrolyte
Anode
C07-COS41xxx-16-05-03-xx-002.eps
Fig. 19:
6
Sensor head, top view
Gold cathode
The sensor consists of the following function units:
• Sensor shaft (Fig. 17, pos. 4)
• Sensor head with Gold cathode and anode (Fig. 18, Fig. 19)
• Membrane cap with electrolyte filling (Fig. 18, pos. 13 and 8)
• Protection basket (Fig. 17, pos. 5)
!
Endress+Hauser
Note!
• Alternatively to the protection basket, you can use a spray head COR 3 (optional, see
"Accessories") for use in immersed operation with cleaning function.
• The membrane cap screwed onto the sensor head is filled with electrolyte. The screw connection
seals it from the medium.
• The membrane which is in contact with the medium is pretensioned in the factory.
15
Operation
Oxymax W COS41
5.2
Measuring principle
5.2.1
Polarisation
When the sensor is connected to the transmitter, a fixed external voltage is applied between the
cathode and anode. The resulting polarisation current is indicated on a display on the transmitter.
The current starts high but then drops over time. The sensor can only be calibrated when the display
is stable.
5.2.2
Membrane
The oxygen dissolved in the medium is conveyed to the membrane by the incoming flow. The
membrane is only permeable for dissolved gases. Other substances dissolved in the liquid phase e.g.
ionic substances, will not penetrate through the membrane. Therefore, medium conductivity has
no impact on the measuring signal.
5.2.3
Amperometric principle
The oxygen molecules diffused through the membrane are reduced to hydroxide ions (OH-) at the
gold cathode. Silver is oxidised to silver ions (Ag+) at the anode (this forms a silver bromide layer,
AgBr).
A current flows due to the connected electrode release at the gold cathode and accepted at the
anode. In equilibrium, this flow is proportional to the oxygen content of the medium.
This current is converted in the measuring instrument and indicated on the display
as an oxygen concentration in mg/l, as a saturation index in % SAT or as an oxygen
partial pressure in hPa.
5.3
Calibration
Calibration is a means of adapting the transmitter to the characteristic values of the sensor. As no
zero calibration is required for the sensor, a single-point calibration is carried out in the presence of
oxygen.
The sensor requires calibration after:
• first commissioning
• replacing a membrane or electrolyte
• cleaning the gold cathode
• long breaks in operation without power supply
• typical time intervals dependent on operating experience
There are three basic types of calibration:
• in air (preferably saturated water vapour, e.g. near the water surface)
• in air-saturated water
• by entering a reference measured value in the transmitter (sensor remains in the medium).
!
16
Note!
Following only the calibration in air is discribed because it is the easiest and that’s why the
recommended method of calibration.
Calibration in air is only possible if air temperature ≥ – 5 °C.
Endress+Hauser
Oxymax W COS41
Operation
5.3.1
!
Calibration in air
1.
Remove the sensor from the medium.
2.
Clean the outside of the sensor with a damp cloth. Then dry the sensor membrane e.g. by using
a tissue.
3.
If the sensor is removed from a closed pressure system with a process pressure greater than
atmospheric pressure:
– Open the membrane cap to equilibrate the pressure and clean the cap if necessary.
– Replace the electrolyte filling and close the membrane cap again.
– Wait for the polarisation time to end.
4.
Then wait while the sensor adjusts to the temperature of the ambient air. This takes about 20
minutes. Check that the sensor is not in direct sunlight during this time.
5.
If the measured value display on the transmitter is stable, carry out the calibration in
accordance with the Operating Instructions of the transmitter.
6.
Place the sensor in the medium again.
Note!
Make sure you comply with the instructions for calibration in the Operating Instructions of the
transmitter.
5.3.2
Calculation example for the calibration value
As a check, you can calculate the expected calibration value (transmitter display) as shown in the
following example (salinity is 0).
Endress+Hauser
1.
Determine:
– the sensor temperature (ambient air)
– the altitude above sea level
– the current air pressure (=rel. air pressure to sea level) at the time of calibration. (If
undeterminable, use 1013 hPa for an approximate calculation.)
2.
Define:
– the saturation value S acc. to the first table
– the factor K acc. to the second table
°C
S [mg/l]
°C
S [mg/l]
°C
S [mg/l]
°C
S [mg/l]
0
14.64
11
10.99
21
8.90
31
7.42
1
14.23
12
10.75
22
8.73
32
7.30
2
13.83
13
10.51
23
8.57
33
7.18
3
13.45
14
10.28
24
8.41
34
7.06
4
13.09
15
10.06
25
8.25
35
6.94
5
12.75
16
9.85
26
8.11
36
6.83
6
12.42
17
9.64
27
7.96
37
6.72
7
12.11
18
9.45
28
7.82
38
6.61
8
11.81
19
9.26
29
7.69
39
6.51
9
11.53
20
9.08
30
7.55
40
6.41
10
11.25
17
Operation
Oxymax W COS41
Altitude
[m]
K
Altitude
[m]
K
Altitude
[m]
K
Altitude
[m]
K
0
1.000
550
0.938
1050
0.885
1550
0.834
50
0.994
600
0.932
1100
0.879
1600
0.830
100
0.988
650
0.927
1150
0.874
1650
0.825
150
0.982
700
0.922
1200
0.869
1700
0.820
200
0.977
750
0.916
1250
0.864
1750
0.815
250
0.971
800
0.911
1300
0.859
1800
0.810
300
0.966
850
0.905
1350
0.854
1850
0.805
350
0.960
900
0.900
1400
0.849
1900
0.801
400
0.954
950
0.895
1450
0.844
1950
0.796
450
0.949
1000
0.890
1500
0.839
2000
0.792
500
0.943
3.
Determine:
– L = rel. air pressure at calibration (1013 hPa if unknown)
– M = 1.02 for calibration in air resp. 1.00 for calibration in air-saturated water
4.
Calculate the calibration value C:
C=S.K.L.M
Example
• Air calibration at 18°C, altitude 500 m above sea level, air pressure 1022 hPa
• S = 9.45 mg/l, K = 0.943, L = 1.0089, M = 1.02
Calibration value C = 9.17 mg/l.
18
Endress+Hauser
Oxymax W COS41
Commissioning
6
Commissioning
6.1
Function check
Before first commissioning, check if:
• the sensor is correctly installed
• the electrical connection is correct.
If using an assembly with automatic cleaning, check the correct water connection at the assembly
rinse connection.
#
Warning!
Danger of medium leaking off
Before applying compressed air to an assembly with cleaning facility, make sure the connections are
correctly fitted. Otherwise, the assembly may not be insert into the process.
6.2
Polarisation
The sensor was tested in the factory for perfect functionality and is supplied ready for operation.
To prepare for calibration, proceed as follows:
!
"
1.
Remove the sensor protective cap.
2.
Place the externally dry sensor in atmospheric air. The air should be saturated with water
vapour. Therefore, install the sensor as close to the water surface as possible. When calibrating
the sensor membrane, make sure the membrane remains dry. Therefore, avoid any direct
contact with the water surface.
3.
Connect the sensor to the transmitter and switch on the transmitter.
4.
Switch-on the transmitter.
If you connect the sensor to the transmitter Liquisys M COM 223/253-DX/DS, polarisation
is automatically performed after switching on the transmitter.
5.
The polarisation time takes about 1 hour.
Note!
Polarisation starts high, then drops gradually. You will recognise the end of polarisation when the
display stabilises and remains practically constant.
Caution!
• When you remove the sensor from the medium, protect the sensor from strong sunlight.
• Make sure you comply with the instructions for commissioning and calibration in the Operating
Instructions of the transmitter.
6.3
Calibrate
Calibrate the sensor (Calibration in air, → Page 17) immediately after it’s polarisation.
The calibration intervals depend heavily on:
• The application and
• The installation position of the sensor.
The following methods help you determine how long the calibration intervals should be:
1.
Check the sensor one month after its being put into operation by taking it out of the fluid,
drying it and then measuring the oxygen saturation index at air after 10 minutes.
Decide using the results:
a. If the measured value is not at 102 ±2 %SAT, you have to calibrate the sensor.
b. Otherwise, double the length of time to the next inspection.
2.
!
Endress+Hauser
Proceed as per Point 1 after two, four and/or eight months. In this way, you can determine
the optimum calibration interval for your sensor.
Note!
Be sure to calibrate the sensor at least once a year.
19
Maintenance
Oxymax W COS41
7
Maintenance
Maintenance work must carried out at regular intervals. To ensure that it is carried out, we
recommend you enter the maintenance dates into an operations logbook or in an
operations calendar in advance.
The following activities must be carried out:
• Cleaning the sensor
(In particular when the membrane is soiled)
• Check the measuring function:
1. Remove the sensor from the medium.
2. Clean and dry the membrane.
3. After about 10 minutes, measure the oxygen saturation index in air (without recalibration).
4. The measured value should be near to 102% SAT (display of O2 saturation with COM 2x3 by
pressing 4 times the O key).
• Replace a defective membrane or one which cannot be cleaned any more.
• Recalibration.
!
Note!
For regular automatic sensor cleaning, we recommend equipping the measuring point with a
fully-automatic cleaning system, e.g. Chemoclean (see "Accessories").
7.1
Cleaning the sensor
The measurement can be corrupted by sensor fouling or malfunction, e.g.:
• Coatings on the sensor membrane
➠ cause longer response times and a reduced slope.
• Soiling or poisoning of the electrolyte
➠ causes longer response times and false measurement.
To ensure reliable measurement, the sensor must be cleaned at regular intervals. The frequency and
intensity of the cleaning operation depend on the measuring medium.
Clean the sensor:
• before every calibration
• at regular intervals during operation as necessary
• before returning it for repairs.
Depending on the type of soiling, proceed as follows:
"
20
Type of soiling
Cleaning
Salt deposits
Immerse the sensor in drinking water or in 1-5% hydrochloric acid
for a few minutes. Afterwards, rinse it with copious amounts of water.
Dirt particles on the sensor body
(not on the membrane!)
Clean the sensor body mechanically with water and a suitable brush.
Dirt particles on the membrane cap or the
membrane
Clean the membrane with water and a soft sponge.
Caution!
After cleaning, rinse the sensor with copious amounts of clean water.
Endress+Hauser
Oxymax W COS41
Maintenance
7.2
Regenerating the sensor
Parts of the sensor will suffer wear and tear during operation.
Suitable action can restore normal operating functionality. This action includes:
#
Action
Cause
Cleaning the gold cathode
soiled or silver-plated gold cathode
Replacing the sealing ring
visual damage to the sealing ring
Replacing the electrolyte
unstable or implausible measuring signal or electrolyte soiling
Replacing the membrane cap
uncleanable membrane, damaged membrane (hole or overstretch)
Warning!
Before beginning regeneration, switch off the power supply at the transmitter.
7.2.1
Cleaning the gold cathode
The gold cathode only needs to be cleaned when it is visibly soiled or there is a coating of silver on it.
To clean it, proceed as follows:
"
1.
Unscrew the membrane cap from the membrane body.
2.
Carefully clean the gold surface in two stages with the abrasive sheet (contained in scope of
supply) until the (silver) coating is fully removed. Use the green sheet first and then the yellow
sheet.
3.
Clean the electrode with drinking or distilled water.
4.
Fill the membrane cap with fresh electrolyte COY 3-F and screw it back onto the membrane
body (up to the stop).
Caution!
The anode is covered with a silver bromide layer at the factory. Do not clean the anode under
any circumstances!
If the coating of the anode is removed as a result of operation, the sensor is unusable and must be
sent in for recoating. In this case, contact your sales office.
7.2.2
Replacing the sealing ring
Replacing the sealing ring is only necessary when it is visibly damaged. For replacement, use only
the supplied sealing rings COY 31-OR.
7.2.3
Replacing the electrolyte
The electrolyte COY 3-F is slowly used up during measuring operations. The cause of this is
electrochemical substance reactions. No substance reactions occur in de-energised state and the
electrolyte is not used up.
The theoretical life of an electrolyte filling for use in air-saturated drinking water at 20°C is max. 5
years. The electrolyte life is shortened by diffused, dissolved gases such as H2S, NH3 or high
concentrations of CO2.
Particular loads occur with:
• anaerobic stages (e.g. denitrification)
• strongly polluted industrial wastewater, particularly at high temperatures.
Endress+Hauser
21
Maintenance
Oxymax W COS41
#
Warning!
Risk of acid burns!
The electrolyte is strongly alkaline. You must follow the appropriate occupational safety regulations.
Always wear protective gloves and goggles with handling electrolytes.
To replace the electrolyte, proceed as follows:
1.
Remove the membrane cap.
2.
Replace the electrolyte and, if necessary, the membrane cap.
3.
Place the membrane cap back on the membrane body and screw the cap to the stop.
7.2.4
"
Replacing the membrane cap
Caution!
Please, use for the COS 41 sensor only the COY 31-WP (colour: black) membrane cap!
Removing the old membrane cap
1.
Remove the sensor from the medium.
2.
Unscrew the protection basket.
3.
Clean the outside of the sensor.
4.
Unscrew the membrane cap.
5.
If necessary, clean the gold cathode or replace the sealing ring if it is damaged.
6.
Rinse the electrode holder with drinking water.
Installing the new membrane cap
7.
Make sure that there are no dirt particles on the sealing surface.
8.
Fill the complete contents of a plastic ampoule (containing electrolyte COY3-F) into the
membrane cap.
9.
Remove all the air bubbles in the electrolyte by tapping the side of the membrane cap (e.g. with
a pencil).
10. Hold the sensor body at an angle and carefully screw the membrane cap onto it down to the
stop.
11. Screw the protection basket back on.
!
22
Note!
After replacing the membrane cap, polarise and recalibrate the sensor. Then insert the sensor into
the medium and check that no alarm is displayed on the transmitter.
Endress+Hauser
Oxymax W COS41
Accessories
8
Accessories
8.1
Connection accessories
❑Junction box VBM
for cable extension, with 10 terminals, IP 65 / NEMA 4X
Cable entry Pg 13.5
Cable entry NPT ½"
Order no. 50003987
Order no. 51500177
❑Measuring cable CYK 71
for use as extension cable between junction box VBM and transmitter, sold by the metre;
order no. 50085333
❑Measuring cable with TOP 68 connector
COK 41; length: 7 m (22.97 ft); order no. 51506817
COK 41; length: 15 m (49.22 ft); order no. 51506818
8.2
Installation accessories
❑Immersion assembly COA110
for sensor immersion in the basin, PVC pipe resp. PUR floating body with SS 1.4571 (AISI 316Ti)
immersion tube;
Ordering acc. to product structure (see Technical Information TI 035C/07/en, 50052566)
❑Flow assembly COA250
for sensor installation in pipe lines, PVC;
(Technical Information TI111C/07/en, order no. 50068518)
Order no. COA 250-A
❑Immersion assembly Dipfit W CYA611
for sensor immersion in basins, open channels and tanks, PVC;
Ordering acc. to product structure (Technical Information TI 166C/07/en)
❑Immersion assembly CYY 105
for sensor immersion in basins, SS 1.4404 (AISI 316L) pipe, SS 1.4571 (AISI 316Ti) fitting;
Ordering acc. to product structure (Technical Information TI 092C/07/en, 50061228)
❑Basin rim holder CYY 106
for sensor immersion into the basin, SS 1.4301 (AISI 304);
Order no. CYY 106-A
❑Baffle plate OP
extra protection for extreme current conditions;
order no. 50028712
❑Membrane protection basket COY 3-SK
or sensor use in fish ponds;
order no. 50081787
8.3
Measurement, controlling and sensor cleaning
❑Liquisys M COM 223/253-DX/DS
Transmitter with integrated sensor function monitoring, measured value monitoring, free
configuration alarm contact, field or control panel installation, HART® or PROFIBUS possible;
Ordering acc. to product structure, see Technical Information TI 199C/07/en;
order no. 51500281
❑ChemoClean
Injector CYR 10 and programme sequencer CYR 20
Ordering by product structure, see Technical Information (TI 046C/07/en)
❑Spray head COR 3
for sensor cleaning in immersion operation;
order no.: COR 3-0
Endress+Hauser
23
Trouble-shooting
Oxymax W COS41
9
Trouble-shooting
9.1
Trouble-shooting instructions
If any of the following problems occur, test the measuring device as indicated.
Problem
No display, no sensor
reaction
Check
Remedial action
Mains voltage to the transmitter?
Connect mains voltage.
Sensor connected correctly?
Set up correct connection.
Medium flow available?
Create flow.
Coating on the membrane?
Clean the sensor.
Electrolyte in the measuring chamber?
Fill with electrolyte or replace electrolyte.
With a TOP 68 connection: Humidity or dirt in Cleaning of the TOP 68 plug-in
plug ?
connection by using cleaning alcohol.
Displayed value too high
Polarisation complete?
Wait until polarisation time ends.
Last calibration with different sensor?
Recalibrate
Temperature display clearly too low?
Check sensor, if necessary send
sensor in for repair.
Membrane visibly stretched?
Replace membrane cap.
Electrolyte soiled?
Replace electrolyte.
Open sensor. Dry electrodes transmitter
display now at 0?
Check electrical connection. If the
problem still occurs, send the sensor in.
Anode coating dissolved, is the anode silver
instead of brown?
Send in the sensor for recoating.
Gold cathode silver-plated?
Clean the gold cathode.
With a TOP 68 connection: Humidity or dirt in Cleaning of the TOP 68 plug-in
plug ?
connection by using cleaning alcohol.
Displayed value too low
Strong deviations in
displayed value
!
24
Sensor calibrated?
Recalibrate
Medium flow available?
Create flow.
Displayesd temperature clearly too high?
Check sensor, if necessary send
sensor in for repair.
Coating on the membrane?
Clean membrane or replace membrane
cap.
Electrolyte soiled?
Replace electrolyte.
Membrane visibly stretched?
Replace membrane cap.
Open sensor. Dry electrodes transmitter
display now at 0?
Check electrical connection. If the
problem still occurs, send the sensor in.
EMC interference on the measuring system?
Remove outer screening of sensor and
extension cable at terminal S. Cut
measuring and signalling lines from h.v.
power lines.
Note!
Make sure you comply with the instructions for troubleshooting in the Operating Instructions of the
transmitter. If necessary, carry out a test of the transmitter.
Endress+Hauser
Oxymax W COS41
Trouble-shooting
9.2
"
Sensor checks
Caution!
Only authorised and trained personnel may test the sensor.
You will also require a multimeter (voltage, resistance).
Check
Measure
Setpoint
Voltage inspection
With the sensor connected, test the
polarisation voltage on the Liquisys M COM
223/253-DX/DS transmitter
between terminals 90 and 91:
–650 mV
Slope inspection
Place the sensor in the air, and dry with a
paper towel.
After 10 minutes:
approx. 102% SAT (4 times O)
Immerse the sensor in zero solution1.
Display near to 0 mg/l (0% Sat)
Zero point inspection
Open the measuring chamber and dry the
electrodes.
Disconnect the sensor and measure the
resistance between red and brown flexible
wires.
depending on temperature:
5 °C: 74.4 kΩ
10 °C: 58.7 kΩ
15 °C: 46.7 kΩ
20 °C: 37.3 kΩ
25 °C: 30.0 kΩ
30 °C: 24.3 kΩ
Disconnect, open and dry the sensor.
Measure the resistance between gold
cathode and white braided wire. Measure
the resistance between anode and green
braided wire.
< 1 Ω each
Temperature sensor check
Cable check
1
How to use the zero solution:
1.
Fill a large beaker (1.5 - 2 l) with approx. 1 l of water.
2.
Pour a cap-full of the zero solution into the water.
3.
Immerse the sensor into the water and wait a sufficient period of time (15 min. for oxygen
depletion).
The display drops to around 0 mg/l (0 %SAT).
Depending on the conditions (contact surface water/air), the zero solution is stable for up to 12
hours.
!
Endress+Hauser
Note!
If there are deviations from the reference values, follow the troubleshooting instructions or contact
your sales office.
25
Trouble-shooting
Oxymax W COS41
9.3
Spare parts
1
Position
Spare parts kit
order no.
1
Sensor
see product
structure
2
Sealing ring COY 31-OR
– Material: Viton
– 3 pieces
51506985
3
Membrane cap
– Replacement cartridge COY 31-WP for normal response
time
– 2 preterminated replacement cartridges with pretensioned
membrane
51506976
4
Membrane protection basket
2
Electrolyte COY 3-F
without fig. – 10 plastic ampoules, transparent
3
4
2-4
C07-COS41xxx-09-05-06-xx-001.eps
Fig. 20:
9.4
COS 41
50053349
Polishing sheets COY 3-PF
– for gold cathode cleaning
– 10 pieces
51506973
Zero solution
– 3 units to produce 3 x 1 litre oxygen free solution
50001041
Accessories kit, one of each:
– Electrolyte COY 3F
– Replacement cartridge COY 31-WP for normal response
time
– Sealing ring COY 31-OR
– Polishing sheet COY 3-PF
51506784
Return
If the device requires repair, please send it cleaned to the sales centre responsible. Please use the
original packaging, if possible.
Please enclose the completed "Declaration of contamination" (copy the second last page of these
Operating Instructions) with the packaging and the transportation documents.
No repair without completed "Declaration of contamination"!
9.5
Disposal
The device contains electronic components and must therefore be disposed of in accordance with
regulations on the disposal of electronic waste.
Please observe local regulations.
26
Endress+Hauser
Oxymax W COS41
Technical data
10
Technical data
10.1
Input
Measured value
dissolved oxygen [mg/l / % SAT / hPa]
Measuring range
with Liquisys M COM 223/253-DX/DS:
0.05 ... 20.00 mg/l (ppm)
0.00 ... 200 % SAT
0 ... 400 hPa
10.2
Response time
T90: 3 minutes
T99: 9 minutes (each at 20 °C)
Slope
approx. 300 nA (at 20 °C, 1013 hPa)
Polarisation time
< 60 minutes
Minimum flow rate
typical 0.5 cm/s for 95 % measured value display
Drift
with permanent polarisation: < 1 % per month
Zero current
zero current free
Sensor monitoring
with Liquisys M COM 223/253-DX/DS:
cable interruption or short-circuit
incorrect measurement and sensor passivation
10.3
Environment
Storage temperature
filled with electrolyte: –5 ... 50 °C (23 ... 122 °F)
without electrolyte: –20 ... 60 °C (–4 ... 140 °F)
Ingress protection
IP 68
10.4
Endress+Hauser
Performance characteristics
Process
Process pressure
max. 10 bar (145 psi) permissible overpressure
Underpressure operation is not permissible.
Process temperature
–5 ... 50 °C (23 ... 122 °F)
27
Technical data
Oxymax W COS41
10.5
28
Mechanical construction
Design, dimensions
see chapter "Installation"
Weights
with cable length 7 m (22.97 ft): 0.7 kg (1.5 lb.)
with cable length 15 m (49.22 ft): 1.1 kg (2.4 lb.)
with TOP68 plug-in connection: 0.3 kg (0.7 lb.)
Materials
Sensor shaft:
Membrane cap:
Cathode:
Anode:
Process connections
G1 and NPT ¾"
Cable entry
Fixed cable or TOP68 plug-in connection
Cable length incl. extension cable
max. 50 m (164.05 ft)
Temperature compensation
NTC temperature sensor 30 kΩ at 25 °C (77 °F), 0 ... 50 °C (32 ... 122 °F)
Membrane thickness
approx. 50 µm
POM
POM
Gold
Silver / silver bromide
Endress+Hauser
Oxymax W COS41
Index
A
Accessories
Baffle plate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
For cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
For connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
For installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Measurement, controlling and cleaning . . . . . . . . . . . . 23
Membrane protection basket . . . . . . . . . . . . . . . . . . . . 23
Amperometric principle. . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Angle of installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Angle of . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Measuring point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Place . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
M
Baffle plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Basin rim holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Measuring point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Mechanical construction . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Membrane. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Membrane protection basket . . . . . . . . . . . . . . . . . . . . . . . 23
C
O
Cable connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Cable length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Calculating the calibration value. . . . . . . . . . . . . . . . . . 17
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
In air . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Chain assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Checking
Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Cleaning
Gold cathode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 19
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
B
D
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Drift . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
E
Electrical connection
Direct connection . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Via junction box . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
13
13
27
24
F
Floating body . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Flow assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Flow operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
I
Icons. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Immersion assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Immersion operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Incoming acceptance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Input. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 7–8
Endress+Hauser
P
Performance characteristics . . . . . . . . . . . . . . . . . . . . . . . . 27
Polarisation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16, 19
Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Product structure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
R
Regenerating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Replacing
Electrolyte . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Membrane cap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Sealing ring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Return. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 26
S
Safety icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Sealing ring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Sensor
Checks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Regenerating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Spare parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Storage temperature. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
T
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Mechanical construction . . . . . . . . . . . . . . . . . . . . . . . 28
Performance characteristics . . . . . . . . . . . . . . . . . . . . . 27
Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Transport. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
29
Oxymax W COS41
U
Upright post . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
W
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Z
Zero current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
30
Endress+Hauser
Declaration of contamination
Dear customer,
Because of legal determinations and for the safety of our employees and operating equipment, we need this "Declaration of
contamination" with your signature before your order can be handled. Please, include the completely filled in declaration with the
device and the shipping documents in any case. Add also safety sheets and / or specific handling instructions if necessary.
Type of device / sensor:
Serial no.:
Medium / concentration:
Temperature:
Pressure:
Cleaned with:
Conductivity:
Viscosity:
Warning hints for medium used (mark the appropriate hints)
SAFE
radioactive
explosive
caustic
poisonous
harmful to
health
biologically
hazardous
inflammable
safe
Reason for return
Company data
Company:
Contact person:
Department:
Address:
Phone:
Fax / e-mail:
Your order no.:
I hereby certify that the returned equipment has been cleaned and decontaminated acc. to good industrial practices and is in
compliance with all regulations. This equipment poses no health or safety risks due to contamination.
(Place, date)
More information about services and repairs:
www.services.endress.com
(Company stamp and legally binding signature)
www.endress.com/worldwide
BA284C/07/en/04.04
Printed in Germany / FM+SGML 6.0 / DT
51506691