Det-Tronics Infiniti® Gas Transmitter U9500 Catalytic Combustible Gas (CGS) Detector Instruction Manual
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Instructions
95-8444
Infiniti ® Gas Transmitter
U9500
10.1
Rev: 4/12
95-8444
Table Of Contents
APPLICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
SETUP PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . 20
FEATURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
IMPORTANT SAFETY NOTES . . . . . . . . . . . . . . . . . 4
INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Wiring Requirements . . . . . . . . . . . . . . . . . . . . . . 4
Wiring Procedure . . . . . . . . . . . . . . . . . . . . . . . . 4
Typical System Wiring . . . . . . . . . . . . . . . . . . . . . 4
Controller/Signal Receiver Options . . . . . . . . . . . 9
Calibration Recommendations . . . . . . . . . . . . . 20
Auto Calibration Procedure . . . . . . . . . . . . . . . . 24
Manual Calibration Procedure . . . . . . . . . . . . . 26
Calibration Procedure for Infiniti with C7065E
Oxygen Sensor . . . . . . . . . . . . . . . . . . . . . . 26
CGS Sensor Replacement . . . . . . . . . . . . . . . . 27
TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . 29
MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Sensor and Hydrophobic Filter Replacement . . 31
Sensor Separation . . . . . . . . . . . . . . . . . . . . . . . 9
SPARE PARTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
DISPLAY AND CONTROLS, OPTIONS, DEFAULTS . . 12
DEVICE REPAIR AND RETURN . . . . . . . . . . . . . . . 33
Display and Controls . . . . . . . . . . . . . . . . . . . . . 12
Programming Options . . . . . . . . . . . . . . . . . . . . 12
ORDERING INFORMATION . . . . . . . . . . . . . . . . . . 33
Infiniti Sensor Compatibility . . . . . . . . . . . . . . . 15
Infiniti Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . 15
MODES OF OPERATION . . . . . . . . . . . . . . . . . . . . 15
INSTALLATION CHECKLIST . . . . . . . . . . . . . . . . . . 19
STARTUP PROCEDURE . . . . . . . . . . . . . . . . . . . . 19
SETPOINT DISPLAY MODE . . . . . . . . . . . . . . . . . . 19
APPENDIX A — FM APPROVAL AND
PERFORMANCE REPORT . . . . . . . . . . . . . . A-1
APPENDIX B — CSA APPROVAL . . . . . . . . . . . . . B-1
APPENDIX C — ATEX / CE APPROVAL . . . . . . . . C-1
APPENDIX D — IECEx APPROVAL . . . . . . . . . . . D-1
APPENDIX E — ADDITIONAL APPROVALS . . . . . E-1
Note: For information regarding MOS H2S please refer to form: 95-8532
INSTRUCTIONS
Infiniti ® Gas Transmitter
Model U9500
APPLICATION
The Infiniti® Gas Transmitter Model U9500 is designed
for use with Detector Electronics Corporation’s (DetTronics®) gas sensors for detection of hazardous gases.
Gas concentrations are displayed on an alphanumeric
display in %LFL or ppm, depending on the type of
gas being detected. The transmitter provides a user
selectable isolated/non-isolated 4 to 20 milliampere
output signal that corresponds to the detected gas
concentration. An optional relay package is available to
provide contact outputs for alarm and fault conditions.
The electronics of the transmitter is contained in an
explosion-proof housing. The transmitter is used with
a single sensor, which may be either coupled directly
with the U9500 housing, or remotely located using a
sensor termination box. The Infiniti offers non-intrusive
calibration by actuating a magnetic reed switch with a
hand-held magnet. It is not necessary to remove the
housing cover to actuate the magnetic reed switch.
Typical Infiniti applications include:
––
––
––
Confined spaces where gas/vapor leaks can
concentrate to explosive or toxic levels or where
oxygen levels need to be monitored (enclosed
processing areas).
Known high risk leak locations (points within
processing areas, compressor buildings).
General open area coverage (storage areas).
FEATURES
––
––
––
––
––
10.1
Non-intrusive calibration using the internal magnetic
reed switch, or an externally located pushbutton (not
included).
Automatic fault diagnostics and graphic message
annunciation.
Optional relay package provides three alarm relays
(high, low, and auxiliary) and one fault relay.
Models available for use with catalytic combustible
sensor, point infrared combustible, electrochemical
(EC) H2S, and CO, CL2, SO2, and O2 electrochemical
sensors.
Variety of multi-port enclosures available.
©Detector Electronics Corporation 2012
SPECIFICATIONS
INPUT VOLTAGE—
24 Vdc. Operating range is 18 to 32 Vdc including ripple.
POWER CONSUMPTION WITH SENSOR—
See Table 1.
CURRENT OUTPUT—
Linear 4 to 20 mA output.
Maximum loop resistance is 600 ohms at 20 to 32 Vdc.
Selectable isolated or non-isolated operation.
Factory set 2.0 mA output indicates unit is in Calibration
mode (field adjustable).
Less than 1.0 mA output indicates fault condition.
DISPLAY—
Eight character alphanumeric display indicates power
on, gas concentration, alarm and fault conditions. It also
enables field adjustment of alarm setpoints, calibration
gas concentration, full scale measurement range and
gas type to be detected.
Rev: 4/12 95-8444
Table 1—Power Consumption with Sensor
Type
Toxic and Oxygen
Combustible
PointWatchTM PIR9400
PointWatch Eclipse® PIRECL
Infiniti with alarm relays
(sensor connected, 24 Vdc+ input)
Nominal
Maximum
3 Watts
4.6 Watts
8.1 Watts
7.6 Watts
Infiniti without alarm relays
(sensor connected, 24 Vdc+ input)
Nominal
Maximum
5 Watts
6.5 Watts
10.0 Watts
12.4 Watts
2.1 Watts
3.5 Watts
7.0 Watts
6.5 Watts
2.4 Watts
4 Watts
7.5 Watts
9.9 Watts
NOTE: Maximum standalone power consumption is 7.7 Watts.
SHIPPING WEIGHT (Transmitter and Junction Box)—
Aluminum:
4.15 pounds (1.88 kilograms)
Stainless steel: 10.5 pounds (4.76 kilograms).
DETECTION RANGE—
Hydrocarbon:
0 to 100% LFL for PointWatch IR
detector. (FM/CSA verified)
Catalytic:
0 to 100% LFL.
Hydrogen Sulfide 0 to 100 ppm, 0 to 50 ppm, 0 to 20 ppm
(electrochemical) (FM/CSA verified)
ENCLOSURE MATERIAL—
Epoxy coated 356 alloy aluminum or 316 stainless steel.
0 to 10 ppm.
Chlorine:
Oxygen:
0 to 25% by volume.
Carbon Monoxide: 0 to 100 ppm, 0 to 500 ppm, 0 to
1000 ppm.
Sulfur Dioxide:
0 to 100 ppm.
DIMENSIONS—
See Figure 1.
CERTIFICATION—
RELAY CONTACTS (OPTIONAL)—
Three Alarm relays: Form C, 5 amperes at 30 Vdc.
Selectable normally energized or de-energized as a
group. Low, auxiliary and high alarm relay contacts are
selectable as a group for latching or non-latching.
FM
®
APPROVED
Warning
When in non-latching mode, the control device
must latch the alarm output.
For complete approval details, refer to the appropriate
Appendix:
Appendix A - FM
Appendix B - CSA
Appendix C - ATEX/CE
Appendix D - IECEx
Appendix E - Additional approvals.
One Fault relay: Form C, 5 amperes at 30 Vdc. Normally
energized for no fault condition with power applied to
device.
TEMPERATURE RANGE—
–40°F to +167°F (–40°C to +75°C).
Operating:
Storage:
–67°F to +185°F (–55°C to +85°C).
WIRING—
Power: 18 AWG (0.75 mm2) minimum is recommended
for power wiring to the transmitter. Larger diameter wire
may be required to maintain a minimum of 16 Vdc at the
transmitter for all operating conditions for all sensors
except PointWatch. A minimum of 18 Vdc is required
at the transmitter for all operating conditions for the
PointWatch sensor. Maximum wire size for terminals is
12 AWG (4 mm2).
10.1
2
95-8444
5.86
(14.9)
5.2
(13.2)
2.7
(6.9)
6.57
(16.7)
4.7
(11.9)
3.46
(8.8)
1.28
(3.3)
B2307
NOTE: TO MOUNT THE TRANSMITTER, DRILL 5/6" (7.9375mm) HOLES
AND USE 1/4" (6.35mm) BOLTS.
Figure 1—Infiniti Transmitter Dimensions in Inches (Centimeters)
DESCRIPTION
Two additional options are available for resetting the
transmitter. A user supplied switch can be wired to the
transmitter terminal block and used to reset the device
from a remote location.
The U9500 is a single channel gas detection device. In
addition to the standard 4-20 mA analog signal output,
the U9500 offers 4 optional relay outputs for fault and
alarm indications. The 4 outputs are: fault, high alarm,
low alarm, and auxiliary alarm. The relays have form C
(SPDT) contacts. Low, auxiliary and high alarm relay
contacts are selectable as a group for latching or nonlatching operation. During normal operation, the fault
relay is non-latching, but for initialization faults, the fault
relay latches. The alarm relays are also selectable, as a
group, for normally energized or de-energized with no
alarms. The fault relay is always normally energized for
no fault.
In addition, Detector Electronics offers a special optional
enclosure with a pushbutton pre-installed within a multiport electrical enclosure for customers who do not prefer
magnetic switches.
The U9500 is compatible with the entire Detector
Electronics sensor family. In some cases, special
personality cards are required to ensure proper
operation with a specific sensor, including the DetTronics Combustible Gas sensors. The proper U9500
model must be ordered by the customer to ensure
proper operation and markings for the desired sensor
type. Refer to Appendix C for combustible gas sensor
ATEX/CE ratings and installation details. Refer to form
90-1041 for specification details.
NOTE
Since the relays can be programmed to be either
energized or de-energized on alarms, the term
“actuate” is used to indicate that the output is in an
alarm condition.
Modes of operation
The U9500 has three main operating modes:
–– Normal
–– Calibrate
–– Setup (also called Configuration).
Normal mode is the default mode, no buttons pushed.
The U9500 provides an eight character, single line, alphanumeric LCD display. The U9500 has four switches
located on the display. The visible three switches are
single pole, normally open push buttons, and are
designated as: Setup\accept, increase, and decrease.
The fourth switch is a magnetic reed switch, designated
cal/reset, which is used by the customer for non-intrusive
calibrations and for resetting of the U9500.
10.1
Three minor modes also exist:
–– Setpoint display
–– Reset
–– Power up.
3
95-8444
Important safety notes
INSTALLATION
WARNING
Do not apply power to the system with the
enclosure cover removed unless the area has
been de-classified. Do not open the enclosure in a
hazardous area when power is applied.
WIRING REQUIREMENTS
CAUTION
The wiring procedures in this manual are intended
to ensure proper functioning of the device under
normal conditions. However, because of the
many variations in wiring codes and regulations,
total compliance to these ordinances cannot be
guaranteed. Be certain that all wiring complies
with applicable regulations that relate to the
installation of electrical equipment in a hazardous
area. If in doubt, consult a qualified official before
wiring the system.
•
Install per local installation practices and in
accordance with local authority having jurisdiction.
•
See Figures 2 and 3 for wiring size and maximum
distance from the power supply to the transmitter.
IMPORTANT
Proper installation practices must be followed
to ensure that condensation does not enter the
junction box and interfere with the electronic
circuitry. Refer to local installation codes.
WIRING PROCEDURE
The following procedure should be used for installing
and wiring the Infiniti Transmitter.
The Infiniti contains semiconductor devices
that are susceptible to damage by electrostatic
discharge. An electrostatic charge can build
up on the skin and discharge when an object is
touched. Therefore, use caution when handling
the device, taking care not to touch the terminals
or electronic components. Observe the normal
precautions for handling electrostatic sensitive
devices.
Refer to these figures and table when installing and
wiring the Infiniti:
Figure 1 — Infiniti Transmitter Mounting Dimensions
Figure 2 — Transmitter Wiring Requirements for
Toxic and Oxygen Models with Relays
Figure 3 — Transmitter Wiring Requirements for
Combustible Models with Relays
Figure 4 — Field Wiring Terminal with Relay Option
Figure 5 — Field Wiring Terminal without Relay
Option
Figure 6 — Transient Suppression for Inductive
Loads
Table 2 — Wiring Size and Maximum Distance from
Transmitter to Sensor (for applications
where a sensor termination box is used)
To minimize the risk of damage, handle the
transmitter module by the edges only. Do not
touch the circuit board, or electronic components.
The fault detection circuitry does not monitor the
operation of external response equipment or the
external wiring to these devices. It is important
that these devices be checked periodically to
ensure that they are operational.
TYPICAL SYSTEM WIRING
Exercise caution if an over-range reading is
indicated, since a dangerous condition could exist.
Refer to these typical system wiring diagrams for
guidance when installing and wiring the Infiniti:
Figure 7 -— A Typical System, Infiniti Transmitter with
a Combustible Gas Sensor, Isolated
Current Output and Relay Outputs
Figure 8 — A Typical System, Infiniti Transmitter
with a Toxic Gas/Oxygen Sensor, NonIsolated Current Output and Relay
Outputs
Figure 9 — A Typical System, Infiniti Transmitter with
a PointWatch Detector and Non-Isolated
Current Output
Figure 10 — A Typical System, Infiniti Transmitter with
a PIRECL PointWatch Eclipse Detector
and Non-Isolated Current Output
10.1
4
95-8444
14 AWG
16 AWG
12 AWG
18 AWG
32
32
30
30
28
28
POWER SUPPLY VOLTAGE
POWER SUPPLY VOLTAGE
12 AWG
26
24
22
20
14 AWG
16 AWG
18 AWG
26
24
22
20
18
18
16
16
14
14
0
2000
4000
6000
8000
DISTANCE (FEET)
0
10000
2000
4000
6000
8000
DISTANCE (FEET)
A1673
10000
A1672
Figure 2—Transmitter Wiring Requirements
for Toxic and Oxygen Models with Relays
Figure 3—Transmitter Wiring Requirements
for Combustible Gas Model with Relays
1. Install the unit in a location that is best suited to cover
the area to be protected. Whenever practical, the
Infiniti should be placed where it is easily accessible
for calibration. For proper operation, sensors should
be pointing down, except for PointWatch, which must
be mounted horizontally. See Figure 1 for mounting
dimensions.
4. Connect the transmitter enclosure to the field wiring
conduit.
The enclosure should be electrically
connected to earth ground.
5. Attach the sensor to the proper entry on the transmitter
enclosure. It is recommended that a layer of teflon
tape be applied to the sensor threading. Route the
wires through the entry, and tighten the sensor to the
enclosure to ensure an explosion-proof installation.
Do not overtighten.
NOTE
When used with the PointWatch infrared
hydrocarbon gas detector, a 1/4 inch spacer is
required between the enclosure and the mounting
surface to allow adequate room for the detector
and calibration windshield, if required.
IMPORTANT
To ease maintenance, periodically apply
a generous amount of Lubriplate grease (part
number 102868-001) to the internal threads on the
junction box cover using an application tool. Do
not use any silicon based products, since some
materials can cause irreversible damage to the
sensing element in the gas sensor.
2. Remove the transmitter from its packing material and
note the location of the field wiring terminals and (on
the catalytic sensor transmitters only) the location of
a small sensor connection plug on the side opposite
the field wiring terminals.
6. Slide the transmitter module into the bracket,
being careful to route field and sensor wires to the
appropriate location within the enclosure for easy
connection to the transmitter module. Ensure that
the transmitter display and pushbuttons (at the top)
are oriented correctly.
3. Remove the cover from the enclosure. The transmitter
module slides into the mounting bracket within the
enclosure and all field and sensor wiring connects
to this module.
–– For catalytic combustible gas transmitters, the
sensor connects to a small plug on one side of the
transmitter and all field wiring connects to terminals
on the other side.
CAUTION
Wires can become pinched between the cover
and base if they are not properly tucked down
inside the enclosure.
–– For all other gas transmitters, the sensor and field
wiring all connect to the terminal block on one side
of the transmitter module.
10.1
5
95-8444
External relays, solenoids, motors, or other devices
that can cause inductive transients should be transient
suppressed. Place a diode across the coil for dc devices.
See Figure 6.
7. Fasten the transmitter module to the mounting
bracket with the two captive screws located inside
the notch on either side of the top of the transmitter.
Remove the wire clip and set it aside.
8. See Figures 4 and 5 for field wiring terminal
designation.
11. An external reset switch can be wired as shown in
Figures 7, 8, 9 and 10. The use of shielded wire is
recommended for wiring the switch.
9. Connect the power and current output leadwires to
the screw terminals on the plug that is provided. See
Figures 7, 8, 9 and 10 for examples of transmitter
wiring. Connect the shield to earth ground at the
power supply. The panel or field end of the shield
should not be grounded at the transmitter. Contact
the factory for further assistance.
12. Check all field wiring to ensure that the proper
connections have been made. Refer to local codes
to be sure wiring and conduit seals are installed
properly.
Warning
When in non-latching mode, the control device
must latch the alarm output.
10. For Relay models, refer to Figures 7 and 8 when
connecting external loads to the relay outputs. For
models without relays, proceed to step 11.
CAUTION
Wires can become pinched between the cover
and base if they are not properly tucked down
inside the enclosure. The wire clip will prevent this
problem when installed correctly.
IMPORTANT
Direct connection of 120/240 vac to the relay
terminals inside the transmitter enclosure is
not permitted, since switching relay contacts
can induce electrical noise into the electronic
circuitry, possibly resulting in a false alarm or other
system malfunction. If the application requires
that ac powered equipment be controlled by the
transmitter, the use of externally located relays is
required.
FAULT RELAY
NORMALLY CLOSED CONTACT
EXTERNAL
RESET
POINTWATCH
CALIBRATE
–
4 TO 20 MILLIAMPERE
INPUT SIGNAL
POWER
POWER
24 VDC +
+
POWER
24 VDC +
+
POWER
COMMON
**
–
**
POWER
COMMON
OUT
+ 4 TO 20 MILLIAMPERE
OUTPUT SIGNAL
– 4 TO 20 MILLIAMPERE
OUTPUT SIGNAL
POWER
PW
RESET
IN
–
+
CAL
S
*SPARE
S
*SPARE
LOW RELAY
NC COM NO
FAULT RELAY
COMMON CONTACT
AUX RELAY
NC COM NO
FLT RELAY
NO COM NC
FAULT RELAY
NORMALLY OPEN CONTACT
HIGH RELAY
NC COM NO
13. Replace the wire clip and place the cover back on
the transmitter enclosure.
HIGH ALARM RELAY
NORMALLY OPEN CONTACT
HIGH ALARM RELAY
COMMON CONTACT
HIGH ALARM RELAY
NORMALLY CLOSED CONTACT
LOW ALARM RELAY
NORMALLY OPEN CONTACT
LOW ALARM RELAY
COMMON CONTACT
LOW ALARM RELAY
NORMALLY CLOSED CONTACT
AUXILIARY ALARM RELAY
NORMALLY OPEN CONTACT
AUXILIARY ALARM RELAY
COMMON CONTACT
AUXILIARY ALARM RELAY
NORMALLY CLOSED CONTACT
* THE SPARE TERMINALS ARE CONNECTED TOGETHER INTERNALLY
AND ARE NOT CONNECTED TO ANY OTHER TERMINALS INTERNALLY
** THESE TERMINALS ARE CONNECTED INTERNALLY
B1904
FLT RELAY
NO COM NC
Figure 4—Field Wiring Terminals, Transmitter with Relays
**
POWER
COMMON
–
**
POWER
24 VDC +
HOT
POWER
+
+
–
PW
RESET
IN
CAL
POWER
POWER
24 VDC +
S
POWER
COMMON
S
*SPARE
*SPARE
{
+
RELAY
EXTERNAL
120 VOLT – V13OLA10A
RESET
SOLENOIDS
MOV
+ 4 TO 20 MILLIAMPERE MOTORS
POINTWATCH
220/240 VOLT – V275LA20A
OUTPUT SIGNAL
INDUCTION DEVICES
CALIBRATE
– 4 TO 20 MILLIAMPERE
4 TO 20 MILLIAMPERE
NEUTRAL
OUTPUT SIGNAL
INPUT SIGNAL
–
OUT
+
}
GE
TYPICAL
A0179
POSITIVE
1N4004
TYPICAL
–
DCV LOADS
NEGATIVE
B0179
* THE SPARE TERMINALS ARE CONNECTED TOGETHER INTERNALLY
AND ARE NOT CONNECTED TO ANY OTHER TERMINALS INTERNALLY
** THESE TERMINALS ARE CONNECTED INTERNALLY
A1905
Figure 6—Transient Suppression for Inductive Loads
Figure 5—Field Wiring Terminals, Transmitter without Relays
10.1
6
95-8444
HIGH RELAY
NC COM NO
RESET
PW
IN
CAL
–
LOW RELAY
NC COM NO
+
–
OUT
+
–
SEE NOTE 6
+
+
AUXILIARY ALARM RELAY CONTACTS
SEE NOTES 2, 4, 5
–
–
+
LOW ALARM RELAY CONTACTS
SEE NOTES 2, 4, 5
POWER
POWER
24 VDC +
–
S
S
24 VDC +
FLT RELAY
NO COM NC
EXTERNAL
RESET
AUX RELAY
NC COM NO
FAULT RELAY CONTACTS
SEE NOTES 2, 3
+
SEE NOTE 1
4 TO 20 MA
NOTE 1
THE COMBUSTIBLE GAS SENSOR CONNECTS TO A PLUG
ON THE OPPOSITE SIDE OF THE TRANSMITTER FROM THE ELECTRICAL TERMINALS.
NOTE 2
ALL RELAY CONTACT DESIGNATIONS INDICATE RELAY STATE
WITH NO POWER APPLIED — RELAYS DE-ENERGIZED.
NOTE 3
WITH POWER APPLIED AND NO FAULTS DETECTED,
THE FAULT RELAY IS NORMALLY ENERGIZED.
NOTE 4
ALARM RELAYS ARE PROGRAMMABLE AS A GROUP
TO BE EITHER NORMALLY ENERGIZED OR NORMALLY DE-ENERGIZED
WITH NO ALARMS DETECTED. REFER TO THE "SET-UP"SECTION OF THIS MANUAL.
NOTE 5
LOW, AUXILIARY AND HIGH ALARM RELAYS ARE PROGRAMMABLE
AS A GROUP FOR LATCHING OR NON-LATCHING OPERATION.
NOTE 6
THESE TERMINALS ARE CONNECTED INTERNALLY.
WARNING:
WHEN IN NON-LATCHING MODE, THE CONTROL DEVICE MUST LATCH THE ALARM OUTPUT.
COMBUSTIBLE GAS
SENSOR
C1668
HIGH ALARM RELAY CONTACTS
SEE NOTES 2, 4
Figure 7—A Typical System - Infiniti Transmitter with Combustible Gas Sensor, Isolated Current Output and Relay Outputs
–
BLACK
–
OUT
+
+
RED
+
+
–
GREEN
4 TO 20 MA
SENSOR
B1669
POWER
PW
RESET
IN
–
+
CAL
POWER
–
LOW RELAY
NC COM NO
S
S
24 VDC +
SEE NOTE 6
HIGH ALARM RELAY CONTACTS
SEE NOTES 2, 4
LOW ALARM RELAY CONTACTS
SEE NOTES 2, 4, 5
AUX RELAY
NC COM NO
FLT RELAY
NO COM NC
EXTERNAL
RESET
HIGH RELAY
NC COM NO
FAULT RELAY CONTACTS
SEE NOTES 2, 3
AUXILIARY ALARM RELAY CONTACTS
SEE NOTES 2, 4, 5
SEE NOTE 1
NOTE 1
CONNECT THE GREEN SENSOR LEAD TO THE CHASSIS GROUND LUG
ON THE INSIDE BOTTOM OF THE INFINITI ENCLOSURE.
NOTE 2
ALL RELAY CONTACT DESIGNATIONS INDICATE RELAY STATE
WITH NO POWER APPLIED — RELAYS DE-ENERGIZED.
NOTE 3
WITH POWER APPLIED AND NO FAULTS DETECTED,
THE FAULT RELAY IS NORMALLY ENERGIZED.
NOTE 4
ALARM RELAYS ARE PROGRAMMABLE AS A GROUP
TO BE EITHER NORMALLY ENERGIZED OR NORMALLY DE-ENERGIZED
WITH NO ALARMS DETECTED. REFER TO THE "SET-UP"SECTION OF THIS MANUAL.
NOTE 5
LOW, AUXILIARY AND HIGH ALARM RELAYS ARE PROGRAMMABLE
AS A GROUP FOR LATCHING OR NON-LATCHING OPERATION.
NOTE 6
THESE TERMINALS ARE CONNECTED INTERNALLY.
WARNING:
WHEN IN NON-LATCHING MODE, THE CONTROL DEVICE MUST LATCH THE ALARM OUTPUT.
Figure 8—A Typical System - Infiniti Transmitter with a Toxic Gas / Oxygen Sensor, Non-Isolated Current Output and Relay Outputs
10.1
7
95-8444
–
POWER
POWER
PW
RESET
IN
CAL
+
+
+
–
–
OUT
+
+
–
–
HIGH RELAY
NC COM NO
LOW RELAY
NC COM NO
S
S
24 VDC +
AUX RELAY
NC COM NO
FLT RELAY
NO COM NC
EXTERNAL
RESET
4 TO 20 MA
BLACK
RED
YELLOW
WHITE
GREEN
B1670
CONNECT THE GREEN SENSOR LEAD TO THE CHASSIS GROUND LUG
ON THE INSIDE BOTTOM OF THE INFINITI ENCLOSURE
POINTWATCH DETECTOR
POWER
PW
RESET
IN
–
+
CAL
–
POWER
+
OUT
+
–
+
LOW RELAY
NC COM NO
S
S
–
AUX RELAY
NC COM NO
FLT RELAY
NO COM NC
EXTERNAL
RESET
HIGH RELAY
NC COM NO
Figure 9—A Typical System - Infiniti Transmitter with a PointWatch Detector and Non-Isolated Current Output
4 TO 20 MA
24 VDC +
+
–
WIRING TO OPTIONAL
RELAY BOARD
NO USER CONNECTION
24 VDC–
1
24 VDC+
2
CALIBRATE
3
24 VDC–
4
24 VDC+
5
+ 4-20 MA
6
– 4-20 MA
7
RS-485 B
8
RS-485 A
9
RELAY POWER (RED)
10
FAULT (ORANGE)
11
LOW ALARM (WHITE)
12
HIGH ALARM (YELLOW)
13
1
A2168
NOTES: 1 250 OHM RESISTOR REQUIRED IF USING HART PORT COMMUNICATIONS.
Figure 10—A Typical System, Infiniti Transmitter with a PIRECL PointWatch Eclipse Detector and Non-Isolated Current Output
10.1
8
95-8444
CONTROLLER/SIGNAL RECEIVER OPTIONS
SENSOR SEPARATION
All Model U9500 Series transmitters offer an analog 4
to 20 mA output, which is configurable for electrically
isolated or non-isolated operation. The analog signal is
capable of driving a maximum 600 ohm resistance load.
Typical controllers used with the U9500 Series include
the family of R8471 Controllers, or the Model 8000 eight
channel controller for % LFL flammable gas monitoring.
It is also acceptable to use the U9500 with other devices
or systems capable of accepting 4 to 20 mA signal
inputs. The following illustrations provide examples
of recommended interconnecting wiring between the
U9500 and analog signal receivers:
Det-Tronics sensor termination boxes (STBs) enable
the installation of the sensor separately from the Infiniti
transmitter module. Three conductor shielded cable must
be used to prevent possible nuisance EMI/RFI. Contact
the factory for ordering information. (See Appendix C for
CE/ATEX rating details.) Refer to these typical system
wiring diagrams for guidance:
Table 2 — Wiring Size and Maximum Distance from
Transmitter to Sensor
Figure 14 — Sensor Separation with Combustible Gas
Sensor and Infiniti Transmitter
Figure 15 — Sensor Separation with PointWatch and
Infiniti Transmitter
Figure 16 — Sensor Separation with Toxic and
Oxygen Gas Sensors and Infiniti
Transmitter
Figure 17 — Sensor Separation with Chlorine Gas
Sensor, Intrinsically Safe barrier and
Infiniti Transmitter
Figure 11 — A Typical System, Infiniti Transmitter with
PointWatch Wired to R8471 Controller,
Non-Isolated Current Output
Figure 12 — A Typical System, Infiniti Transmitter
with PointWatch Wired to a Model 8000
Controller, Isolated Current Output
Figure 13 — A Typical System, Infiniti Transmitter
used with PLC Analog Input Module,
Isolated/Non-Isolated Current Output
R8471 CONTROLLER
–
+
CHASSIS GROUND
2
*
–
4-20 MA
3
5
+
6
SIGNAL
–
7
HIGH ALARM / OC
10
AUX. ALARM
11
AUX. ALARM / OC
12
LOW ALARM
13
LOW ALARM / OC
14
FAULT
15
FAULT / OC
16
RESET
HIGH
ALARM
AUXILIARY
ALARM
POWER
PW
RESET
IN
– +
CAL
9
POWER
OUT
–
+
+
–
8
HIGH ALARM
S
EXTERNAL RESET
*
S
–
POWER
HIGH RELAY
NC COM NO
4
GROUND
LOW RELAY
NC COM NO
18 TO 32VDC +
SENSOR
+
FLT RELAY
NO COM NC
POWER
24 VDC
1
AUX RELAY
NC COM NO
ISOLATED OUTPUT
CURRENT LOOP
A2034
BLACK
RED
YELLOW
WHITE
GREEN
LOW
ALARM
POINTWATCH DETECTOR
FAULT
*NO CONNECTION
Figure 11—A Typical System, Infiniti Transmitter with PointWatch Wired to R8471 Controller, Non-Isolated Current Output
10.1
9
95-8444
MODEL 8000 CONTROLLER
TB 2
TB 1
1
2
U9500 TRANSMITTER NO. 1
3
LOW ALARM
–
TROUBLE
TB 3
TB 4
TB 5
S
24 VDC
BATTERY
+
3 CTR
6
3–
6
3+
6
2 CTR
7
2–
7
2+
7
1 CTR
8
1–
8
1+
8
–
+
POWER
PW
RESET
IN
– +
CAL
24 VDC +
**
–
POWER
OUT
–
+
+
–
FLOAT
TYPE
CHARGER
S
+
LOW RELAY
NC COM NO
FLT RELAY
NO COM NC
5
HIGH RELAY
NC COM NO
HIGH ALARM
4
AUX RELAY
NC COM NO
CONNECTION
NOT RECOMMENDED
BLACK
RED
YELLOW
WHITE
GREEN
POINTWATCH DETECTOR
NO. 2*
NO. 3*
B2035
* IDENTICAL TO TRANSMITTER NO. 1 WIRING.
** INSTALL 250 OHM, 12 WATT RESISTOR ACROSS (+) AND (–) TERMINALS FOR ALL CHANNELS SELECTED FOR SERVICE.
THIS RESISTOR WILL SATISFY POWER MONITORING REQUIREMENTS (ONLY CHANNEL 3 SHOWN FOR CLARITY).
S
INPUT
+
ISOLATED CURRENT OUTPUT
INPUT
NON-ISOLATED CURRENT OUTPUT
24 VDC
OUTPUT
S
INPUT
HIGH RELAY
NC COM NO
FLT RELAY
NO COM NC
INPUT
RETURN
POWER
OUT
–
+
+
–
–
24 VDC
POWER
S
FLT RELAY
NO COM NC
AUX RELAY
NC COM NO
LOW RELAY
NC COM NO
HIGH RELAY
NC COM NO
POWER
PW
RESET
IN
–
+
CAL
LOW RELAY
NC COM NO
POWER
OUT
–
+
+
–
24 VDC
OUTPUT
S
AUX RELAY
NC COM NO
POWER
PW
RESET
IN
–
+
CAL
Figure 12—A Typical System, Infiniti Transmitter with PointWatch Wired to a Model 8000 Controller, Isolated Current Output
+
A2036
RETURN
–
24 VDC
POWER
Figure 13—A Typical System, Infiniti Transmitter used with PLC Analog Input Module, Isolated/Non-Isolated Current Output
10.1
10
95-8444
Table 2—Maximum Sensor Separation Distance
Maximum Transmitter to Sensor Distance
Combustible
Toxic and Oxygen
Point Infrared
Wire Size
(mm2)
Wire Size
(AWG)
Feet
Meters
Feet
Meters
Feet
Meters
1.0
18
100
30
12,000
3,600
700
210
1.5
16
150
45
20,000
6,100
1100
335
2.5
14
250
75
32,000
9,700
1800
545
4.0
12
400
120
50,000
15,000
2800
850
INFINITI TRANSMITTER
ATTACH CABLE SHIELD
TO GND LUG
GND
CONNECT TO TRANSMITTER
SENSOR TERMINATION BOX
WHT
BLK
BLK
WHT
RED
RED
PIGTAIL CONNECTOR
(PROVIDED IN SENSOR
TERMINATION BOX)
NOTES
• SHIELDED CABLE REQUIRED.
• GROUND CABLE SHIELD AT TRANSMITTER END ONLY.
2309A
CONNECT CABLE LEADWIRES
FROM SENSOR TERMINATION BOX
CONNECT KEYED SENSOR
PLUG TO PIN CONNECTOR
Figure 14—Sensor Separation with Combustible Gas Sensor and Infiniti Transmitter
POWER
PW
RESET
CAL IN
– +
OUT
+
– + + –
–
POWER
–
+
LOW RELAY HIGH RELAY
S
S
24 VDC +
POWER SUPPLY
AUX RELAY
NO COM NC
FAULT RELAY
INTERNAL GROUND SCREW
NO COM NC NO COM NC NO COM NC
INFINITI TRANSMITTER JUNCTION BOX
4 TO 20 MILLIAMPERES
DET-TRONICS JUNCTION BOX
INTERNAL
GROUND SCREW
SPARE
CAL
4 – 20
CAL
4 – 20
RET
RET
+24
+24
GREEN
YELLOW
WHITE
BLACK
RED
POINTWATCH
DETECTOR
B1907
Figure 15—Sensor Separation with PointWatch and Infiniti Transmitter
10.1
11
95-8444
4 TO 20 MILLIAMPERES IN
LOW RELAY
INFINITI TRANSMITTER OR
R8471 CONTROLLER
SEE NOTE 1
AUX RELAY
LOW RELAY HIGH RELAY
POWER
PW
RESET
CAL IN
– +
AUX RELAY
S
–
+
– + + –
POWER
PW
RESET
CAL IN
– +
+
OUT
– + + –
–
POWER
–
+
24 VDC +
POWER SUPPLY
POWER OUT
S
S
24 VDC +
POWER SUPPLY
S
NO COM NC
FAULT RELAY
GROUND SCREW
NO COM NC NO COM NC NO COM NC
INFINITI TRANSMITTER JUNCTION BOX
SENSOR POWER (+)
4 TO 20 MILLIAMPERES
SENSOR TERMINATION BOX
+
+
D
RE
K
BL
2
TWO CONDUCTOR
SHIELDED CABLE
(CONNECT AT THE
TRANSMITTER
JUNCTION BOX ONLY)
MTL 787S
I.S. BARRIER
4
G
HAZARDOUS AREA
SENSOR TERMINATION BOX
* CONNECT THE GREEN
SENSOR LEAD TO THE
CHASSIS GROUND LUG
IN THE INSIDE BOTTOM
OF THE JUNCTION BOX.
–
+
NO CONNECTION
+
GROUND LUG*
–
G
ED
R
–
LK
B
G
TOXIC SENSOR
* CONNECT THE GREEN
SENSOR LEAD TO THE
CHASSIS GROUND LUG
IN THE INSIDE BOTTOM
OF THE JUNCTION BOX.
GROUND LUG*
Figure 16—Sensor Separation with Toxic* and Oxygen Gas Sensors
and Infiniti Transmitter
NO CONNECTION
C7067E
CHLORINE
SENSOR
* Excluding Chlorine sensor, see Figure 17.
NOTES:
1. NEVER EXCEED 25.5 VOLTS DC POWER SUPPLY INPUT VOLTAGE OR
BARRIER DAMAGE WILL RESULT.
2. MAXIMUM TOTAL FIELD WIRING SIGNAL LOOP IMPEDANCE IS
130 OHMS (EXCLUDING BARRIER).
3. INTRINSICALLY SAFE WIRING AND INSTALLATION PRACTICES MUST BE FOLLOWED.
4. INTRINSICALLY SAFE BARRIERS ARE NOT COMPATIBLE WITH “GROUND-FAULT”
MONITORED POWER SUPPLIES.
5. INTRINSICALLY SAFE BARRIER MAY BE FIELD INSTALLED WITHIN AN
B2311
APPROVED JUNCTION BOX.
DISPLAY AND CONTROLS, OPTIONS,
DEFAULTS
DISPLAY AND CONTROLS
Figure 17—Sensor Separation with Chlorine Gas Sensor, Intrinsically
Safe Barrier and Infiniti Transmitter
The Infiniti display contains an eight character display
for identifying system status conditions and sensor
input, a magnetic reed switch for resetting the unit and
entering different operating modes, and pushbuttons for
programming and calibrating the system. See Figure 18
for the location of indicators and pushbuttons and Tables
3 and 4 for descriptions.
PROGRAMMING OPTIONS
Operating Range
The available operating range settings and the
corresponding default setpoints and calibration gases
are shown in Table 5. Note that not all U9500s support
selectable operating range.
External Reset
The external reset input terminal, when grounded
momentarily, normally initiates a relay/display reset only.
See Figure 9. However, if the “EXT” “CAL” option is
programmed “YES” during the Infiniti setup procedure,
the external reset duplicates the magnetic reed switch
(Cal Magnet) and can be used to perform calibration.
10.1
3
SAFE AREA
–
G
B2310
1
IMPORTANT
The range setting of the transmitter must match
the output range of the sensor being used or the
system will fail to operate correctly. For example,
if a carbon monoxide (CO) sensor with a range of
0 to 500 ppm is being used, the range setting for
the transmitter must be 0 to 500 ppm. Refer to the
“Specifications” section for a listing of available
sensors and ranges.
12
95-8444
Table 3—Display And Controls
Display/Indicator
Description
This display is always on and functions as a power indicator.
The display provides a continuous reading of the sensor input in both the Normal and
Calibrate modes.
In the event of a fault, it identifies the nature of the fault with a fault message.
In other operating modes it shows the alarm setpoints and programmed calibration
gas concentration.
Ex %LFL / ppm
A negative zero drift condition is indicated by a minus (–) sign in the left hand digit.
In the event of an over-range condition, the display will continue to track the sensor
output as long as the over-range condition exists. Combustible gas transmitters will
latch the display at the highest detected gas concentration.
*
Asterisk indicates an alarm (any one) has activated since last reset.
Alarm History Indicator
Blank indicates no alarm has energized since last reset.
Solid black square indicates that the high alarm threshold has been exceeded.
n
High Alarm Status Indicator
For relay models, this means that the relay output has activated.
Blank indicates no alarm.
Solid black square indicates that the auxiliary alarm threshold has been exceeded.
n
Auxiliary Alarm Status Indicator
n
Low Alarm
For relay models, this means that the relay output has activated.
Blank indicates no alarm.
Solid black square indicates that the low alarm threshold has been exceeded.
For relay models, this means that the relay output has activated.
Blank indicates no alarm.
INCREASE
ms
SETUP
m
ACCEPT
DECREASE
tm
CAL/RESET
Used to move to next higher setting during system programming.
Used to initiate Setup routine, to accept reading displayed, and cycle to the next
programming step during Setup (system programming).
Used to move to next lower setting during system programming.
Used for non-intrusive calibration and transmitter reset.
Activated by Cal Magnet from outside the Infiniti enclosure.
Table 4—CAL/RESET Switch and Calibration Magnet Function
Duration of
CAL/RESET
Switch Activation
Function
1 Second
Resets the Infiniti transmitter
2 - 3 Seconds
Enters the Setpoint Display Mode
7 Seconds
10.1
Enters the Calibration Mode
13
95-8444
SENSOR INPUT READING
ALARM HISTORY INDICATOR
ALARM STATUS INDICATORS
}
ALARM
INCREASE
HI AUX LO
SETUP
DECREASE
ACCEPT
CAL/RESET
INCREASE PUSHBUTTON
DECREASE PUSHBUTTON
SETUP/ACCEPT PUSHBUTTON
CAL/RESET MAGNETIC REED SWITCH
(ACTIVATED BY HOLDING CAL MAGNET TO THIS SIDE OF TRANSMITTER
APPROXIMATELY TWO INCHES DOWN FROM THE WINDOW)
B2374
Figure 18—Infiniti Transmitter Controls and Indicators
Table 5—Range Select Settings And Corresponding Default Settings
Default Settings ***
Sensor
Type
Electrochemical
Toxic *
Range Limits
Measurement
Range
High
Low
Aux
Cal
Gas
High
Low
Aux
Cal
Gas
0-10 ppm
2
1
2
5
1-9
0.5-5
0.5-9
3-9
0-20 ppm
4
2
4
10
1-18
0.5-10
0.5-18
6-18
0-50 ppm
10
5
10
25
2-45
1-25
1-45
15-45
0-100 ppm
20
10
20
50
10-90
5-50
5-90
30-90
0-200 ppm
40
20
40
100
20-190
5-100
5-190
60-180
0-500 ppm
100
50
100
250
50-450
25-250
25-450
150-450
0-1000 ppm
200
100
200
500
100-900
50-500
50-900
300-900
Catalytic
Combustible
0-100 % LFL (Fixed)
50
20
50
50
10-60
5-50
5-90
30-99
PointWatch
(HC)
0-100 % LFL (Fixed)
50
20
50
50
10-60
5-50
5-90
50
Oxygen **
0-25% (Fixed)
23
18
18
20.9
22-25
16-20.5
5-25
8-23.5
* Changing the Toxic Measurement Range will reset all alarms and the Cal Gas to the default setting for the selected range.
** For oxygen deficient applications, use the LOW and/or AUX alarms.
For oxygen enrichment applications, use the HIGH and/or AUX alarms.
*** Default settings for all Infiniti relay operation is normally de-energized with non-latching contacts (except fault alarm, which is normally
energized with no faults). Default settings for optional External Reset Pushbutton is “No External Cal” enabled.
Default settings for calibration mode (automatic or manual) is automatic mode. Manual calibration mode is recommended when using the
U9500D chlorine or U9500F Sulfur Dioxide Transmitters.
10.1
14
95-8444
Setpoints
External Calibration
The Infiniti Transmitter has three independent alarm
outputs (low, high and auxiliary), with field selectable
setpoints. For both relay and non-relay models, these
setpoints correspond to the HI, AUX, and LO ALARM
indicators on the display.
Calibration procedures are normally initiated and
executed using the Cal Magnet to activate the CAL/
RESET switch. Choose “YES EXT CAL” during the setup
procedure to allow the additional use of the external reset
input terminal for initiating and executing a calibration
procedure. See Figure 9 for switch wiring.
Calibration Gas Concentration
4-20 Milliampere Calibration
The programmed calibration gas concentration is also
displayed and adjustable. This known concentration
(typically one half of the instrument’s full scale detection
range) of the actual gas or vapor that is expected to be
detected should be used to calibrate the system.
Det-Tronics calibration gas should be used to ensure
proper system calibration performance and accuracy.
The 4 to 20 mA current loop is factory calibrated. Perform
the “YES 4-20 CAL” function in the Setup procedure if a
recalibration is required. A current meter must be placed
in the loop to perform this function.
Calibration Current
The current output during Calibration and while in the
Setup mode is factory set for 2 mA. Perform the “SET
CAL CURRENT” function in the Setup procedure if an
output current other than 2 mA is required.
IMPORTANT
The calibration gas concentration setting typically
matches the concentration of the gas used to
perform the calibration (printed on label of
calibration gas cylinder). Refer to the “Calibration”
section of this manual for details.
Refer to the “Setup Procedure” section for complete
instructions.
Latching or Non-latching Relays
INFINITI SENSOR COMPATIBILITY
The Low, Auxiliary and High alarm relays are
programmable as a group for either latching or nonlatching operation. “LATCH” indicates that they will
latch upon alarm and the unit must be reset to clear.
“nonLATCH” indicates that they will automatically clear
when the alarm clears.
Refer to Table 6 for a list of acceptable sensor inputs to
the Infiniti by model.
INFINITI OUTPUTS
Refer to Table 7 for a description of Infiniti standard 4 to
20 mA outputs and relay output programming options.
Alarm Relays Normally Energized or Normally
De-Energized
MODES OF OPERATION
The three alarm relays (High, Low and Auxiliary) are
programmed as a group. “DE-ENERG” indicates that
the three alarm relays will be normally de-energized and
will energize upon alarm. “ENERG” indicates that the
three alarm relays will be normally energized and will deenergize upon alarm.
Refer to Table 8 for a description of Infiniti modes of
operation.
Automatic or Manual Calibration
Manual calibration requires that the operator use the Cal
magnet to activate the CAL/RESET switch, signaling an
acceptance of the present calibration reading. Auto Cal
is described in the “Calibration” section of this manual.
10.1
15
95-8444
Table 6—Infiniti Sensor Compatibility
Infiniti Model
Accepts
Combustible
• U9500A
Accepts inputs from Det-Tronics Catalytic Combustible gas sensor only.
Toxic and Oxygen Models
• U9500B Hydrogen Sulfide
• U9500C Oxygen
• U9500D Chlorine
• U9500E Carbon Monoxide
• U9500F Sulfur Dioxide
Accepts 4 to 20 mA inputs only.
• C7064E Hydrogen Sulfide Sensor
• C7065E Oxygen Sensor
• C7067E Chlorine Sensor *
• C7066E Carbon Monoxide Sensor
• C7068E Sulfur Dioxide Sensor
Hydrocarbon (HC)
• U9500H
Accepts inputs from the Det-Tronics PointWatch infrared hydrocarbon gas detector.
* External intrinsically-safe barrier required for hazardous area approval (I.S. barrier not
available from Det-Tronics).
Table 7—Infiniti Outputs
Output
4 to 20 Milliamperes (Standard)
Description
• The linear 4 to 20 mA output:
– corresponds to the field-programmed gas range.
– can be calibrated in the field for maximum accuracy.
– is user selectable for isolated or non-isolated use.
Refer to the “Setup ” section for programming information.
4 to 20 mA with Relays (Optional)
• Option includes a set of four relays:
–High Alarm
–Low Alarm
–Auxiliary Alarm
–Fault
• The High, Low and Auxiliary Alarm relays are programmable
as a group for either normally energized or normally de-energized
operation.
• Low Alarm, Auxiliary Alarm, and High Alarm relays are programmable as
a group for either latching or non-latching operation.
• Fault relay is normally energized with no faults. Faults are generally
non-latching, but faults that occur during calibration and warmup must be
reset.
• Latching relays are reset using either the Cal magnet or an external reset
switch (field installed, user supplied).
10.1
16
95-8444
Table 8—Infiniti Modes of Operation
Operating Mode Description
• When power is applied to the transmitter, it enters a Warmup mode to allow the sensor output to stabilize
before beginning normal operation. During this time the:
–Fault relay is de-energized.
– Alarm relays go to non-alarm state (either energized or non-energized, as programmed)
– Display alternates between “WARM-UP” and “Ex” or “Toxic” or “PTIR” or “Oxygen” depending on specific
model.
– Current output indicates a fault condition (less than 1.0 mA).
• The transmitter will stay in the Warmup mode for at least six seconds.
Warmup
– If the detected gas at the end of the six second warmup period is higher than the lowest alarm setting or if
faults are present, the transmitter will remain in the Warmup mode until the detected gas falls below the
lowest alarm setting and no faults are present, or five minutes, whichever is shortest.
– If an alarm condition exists at the end of the five minute warmup, the transmitter will enter the Normal
operating mode and the alarm(s) will be annunciated.
– If a fault is present after the five minute warmup, the transmitter will indicate the fault, the fault relay will
remain de-energized, and the current output will be less than 1.0 mA.
• At the end of the warmup period with no faults or alarms present, the transmitter automatically enters the
Normal operating mode (fault relay energizes, alarm relays remain in non-alarm state), and the current output
will rise to 4.0 mA.
• In Normal operating mode with no alarm condition:
– Display is on and indicates the detected gas concentration.
– Alarm relays (on relay models) are in their normal state (energized or de-energized as programmed).
– 4 to 20 mA output signal level corresponds to the detected gas concentration.
– Fault relay (on relay models) is energized.
• In the Normal operating mode with an alarm condition occurring:
– Display indicates the detected gas concentration.
– Low, Auxiliary or High display shows a black square, indicating alarm condition.
– Affected relay changes state (on relay models).
– 4 to 20 mA output signal level corresponds to the detected gas concentration.
– Fault relay output energized (on relay models).
– Relay History Indicator display shows an asterisk to indicate an alarm was activated.
• When the signal decreases below setpoint:
Normal
– Display and 4 to 20 mA output continues to track the detected gas concentration.
– With latching operation programmed, numeric display shows detected gas concentration and low auxilary,
or high display shows a black square indicating latched alarm output.
– With non-latching operation programmed, low, auxiliary and/or high display goes blank and Alarm relay
outputs return to their normal state.
– Asterisk in Alarm History Indicator display stays on to indicate an alarm since last reset.
• In the event of a system fault:
– The normally energized Fault output is de-energized and the corresponding fault message is displayed.
– 4 to 20 mA output drops to less than 1.0 mA.
• In the event of an alarm condition and a system fault both occurring:
– In most cases, the first condition that occurs will be indicated by the current and relay outputs and on the
display.
– The exceptions are “CAL ABORTED” and “SENSOR E.O.L.” faults which both may occur during the
calibration procedure. If an alarm occurs with these faults, the alarm will over-ride the fault and will be
indicated.
10.1
17
95-8444
Table 8—Infiniti Modes of Operation (Continued)
Operating Mode Description
The Reset mode is entered by holding the Cal magnet to the side of the transmitter enclosure next to CAL/RESET
on the display face of the unit.
Reset
• Standard – When the Cal magnet is held there for less than one second, the alarm indicators turn off and all
relay outputs return to their normal condition if no alarms or faults are occurring.
• Forced –
When the Cal magnet is held there for 1 to 2 seconds, the alarm indicators turn off and the relay
outputs return to their normal condition even if an alarm or fault condition still exists.
• Remote – Activating the remote reset switch for less than 2 seconds initiates a forced reset. If “YES
EXT CAL” was selected during the setup procedure, the external reset input initiates the setpoint
display cycle when activated for over two seconds.
When the Cal magnet is held to the side of the transmitter enclosure next to CAL/RESET for more than two
seconds, the transmitter enters the Setpoint Display mode. Once in this mode, the Cal magnet can be removed.
The Setpoint Display sequence will complete automatically. (The external reset input terminal can also be used
to initiate the setpoint display and calibration mode if “YES EXT CAL” was selected during the setup procedure.)
In this mode:
• The display sequentially shows the programmed alarm setpoints and calibration gas concentration.
• Each value is displayed for approximately 1.5 seconds.
Setpoint Display
• After completing the sequence, the transmitter automatically returns to the Normal operating mode if the Cal
magnet is no longer being held to the unit (or the external reset input is not activated).
NOTE
If the CAL/RESET switch (or external reset input) is still activated at the end of this cycle, the transmitter
automatically enters the Calibrate mode. If a calibration is not performed, a CAL ABORTED fault will occur. To
exit the Calibrate mode without performing the calibration, cycle power or wait for the calibration fault message,
then reset the unit.
The Setpoint Display mode is used only for displaying the setpoints. Use the “Setup” mode for changing setpoint
and calibration gas values.
Auto calibration is the default setting and the recommended calibration method for all sensors except for Chlorine
(Cl2) and Sulfur Dioxide (SO2). Auto calibration is required for use with PointWatch.
• Auto Cal is an automatic calibration procedure that requires no adjustments by the operator. The Calibrate
mode is entered by holding the Cal magnet to the CAL/RESET magnetic switch until completion of the “Setpoint
Display” sequence described above (approximately 7 seconds). (The external reset input terminal can also be
used to enter the calibrate mode if “YES EXT CAL” was selected during the Setup procedure. Activate the input
until completion of the “Setpoint Display” sequence.)
• For calibration of the Infiniti with the C7065E O2 sensor, there are two modes of Auto Calibration available:
1. Initial / Sensor Replacement Calibration
2. Routine Calibration
Use Option #1 for new sensor start-up or after replacing a sensor. Use Option #2 for all subsequent calibrations
after Option #1 has been completed.
Auto Calibration
• The transmitter performs the Zero adjustments, then signals the operator when to apply and also when to
remove the calibration gas. Upon completion of a successful calibration, the transmitter automatically returns
to the Normal operating mode.
• If the operator fails to complete the calibration procedure, if an error in calibrating occurs, or if a successful
calibration cannot be completed, the transmitter will automatically return to the Normal mode and continue to
use the previous calibration data (after 10 minutes or when the gas level drops below the lowest alarm setpoint).
A fault indication will be displayed until a reset occurs.
• If the microprocessor determines that the sensor cell is approaching the end of its useful life, the message “SEN
AT EOL” will be indicated on the display. A fault indication will be displayed until a reset occurs.
• While in the Calibrate mode, all transmitter outputs are inhibited and the dc current output goes to a preset
level (adjustable from 0 to 20 mA, with a default value of 2.0 mA). See “Setup” section for complete calibration
procedure.
10.1
18
95-8444
Table 8—Infiniti Modes of Operation (Continued)
Operating Mode
Description
Manual Calibration
Use the manual calibration mode when calibrating the Chlorine (Cl2) and Sulfur Dioxide (SO2) sensors.
• Manual Calibration must be selected during the initial set up of Infiniti.
• Manual calibration is similar to Auto Calibration, except that when the Zero and Span (calibration gas
concentration) readings are displayed and judged stable by the operator, they must be manually accepted
as calibration points by holding the Cal magnet to the CAL/RESET magnetic switch (or activate the external
reset input if “YES EXT CAL” was selected during the Setup procedure.)
Setup
In the setup mode, the range (for some gases), alarm setpoints, calibration gas level, current loop levels,
relay operation (latching/non-latching, energized/de-energized), and calibration mode (auto/manual) are
programmed into the transmitter.
NOTE
When used with PointWatch only 50% LFL gas may be used for span concentration. PointWatch will not read
accurately if other concentrations are used.
INSTALLATION CHECKLIST
STARTUP PROCEDURE
The following checklist is provided as a means of
checking the system to be sure that all phases of system
installation are complete and have been performed
correctly.
1. Remove power from all output devices to prevent
actuation.
2. Apply power to the system. If the proper “type of
sensor” indication does not appear in the display
during warmup, consult the factory.
1. Enclosure is securely mounted and sensor is pointing
down. (PointWatch models must be mounted
horizontally. Refer to the PointWatch manual for
details.)
3. When the Warmup period is completed, perform the
Setup procedure.
4. Perform the Calibration procedure.
2. Ensure that local wiring and installation codes are
met.
5. Restore power to the output devices.
3. Power wiring is installed and power source is
operational.
SETPOINT DISPLAY MODE
4. Wiring to external loads and/or monitoring devices is
properly connected.
In this mode, the display sequentially shows the
programmed alarm setpoints and calibration gas
concentration, then returns to the Normal operating mode
(see Table 9). Remove power from all output devices to
prevent actuation.
5. If a sensor termination box is used, shielded
interconnecting wiring is correctly installed.
6. All cable shields are properly connected.
7. Optional sensor accessories (dust/splash guards,
sample draw devices, etc.) are installed, clean, and
in good condition.
8. O-ring is in good condition and the junction box
cover is tightly installed.
9. Monitoring devices and/or response equipment is
operational.
10.1
19
95-8444
Table 9—Setpoint Display Cycle
Display Indicates
Duration and Description
Normal Mode
To enter the Setpoint Display mode, activate the CAL/RESET magnetic switch for 2 to 3 seconds
(hold the Cal magnet to the side of the transmitter enclosure next to CAL/RESET). If “YES
EXT CAL” was selected during the “Setup Procedure,” the external reset input can also be used to
initiate the Setpoint Display Cycle when activated for 2 to 3 seconds.
Low Alarm Setpoint
1.5 seconds
High Alarm Setpoint
1.5 seconds
Auxiliary Alarm Setpoint
1.5 seconds
1.5 seconds
IMPORTANT
For catalytic sensors only, whenever the type of calibration gas that is used differs from
the gases to be detected, a Conversion K - Factor must be applied to ensure proper
operation. Refer to “Catalytic Sensor Calibration” section in this manual for details.
Calibration Gas Concentration
Normal Mode
Remains in Normal mode until another mode is activated.
SETUP PROCEDURE
3. Press and hold the SETUP/ACCEPT button for 1
second using a small screwdriver, then release. This
initiates the Infiniti Setup Mode. Refer to Table 10
to perform setup. Once a programming option has
been accepted, the transmitter will automatically
cycle to the next option.
1. Determine the required alarm setpoint levels,
calibration gas concentration, and other settings
such as range (Table 5), auto/man calibration,
normally energized/de-energized relays, latching/
non-latching relays.
CALIBRATION
CAUTION
Changing the range selection will cause the
alarm setpoints and the calibration gas settings
to default to the setting shown in Table 5. Catalytic
combustible, PointWatch infrared, and oxygen
sensors have preset ranges that are automatically
selected by the Infiniti and cannot be changed.
CALIBRATION RECOMMENDATIONS
The Infiniti can accommodate automatic or manual
calibration and for many sensors either of these methods
are effective. Certain sensors, however, require either
automatic or manual calibration only. The sensors and
method of calibration required and/or recommended are
listed in Table 11. Refer to this table before proceeding
with calibration.
2. Remove the enclosure cover to access the transmitter
display and controls.
Perform Calibration
CAUTION
While in the Setup Mode, the current loop output
will drop to less than 1 mA. For models without
relays, this is the only remote indication that the
unit is not in the Normal operating mode. For
models with relays, the fault relay indicates a
fault. In either case, the transmitter will remain
in the Setup mode until Setup is complete. The
transmitter must be manually stepped through the
sequence below in order to return to the Normal
operating mode.
10.1
20
•
When a new system is initially put into service.
•
When the sensor is replaced.
•
Periodically to verify proper performance for
combustible, toxic and oxygen detectors - typically
at 90 day intervals, however, this frequency
is dependent upon the requirements of each
application. Refer to the PointWatch manual for
calibration recommendations specific to that
device. Refer to Table 12 for the recommended
calibration frequency for sensors manufactured by
Det-Tronics.
95-8444
Table 10—Infiniti Setup
Display
Left portion displays the
present range setting.
Right portion displays
alternating:
“RNG”
“SET”
Note: Not present for
Hydrocarbon (PointWatch).
Left portion displays the
low alarm setpoint.
Right portion displays
alternating:
“LO”
“SET”
Left portion displays the
high alarm setpoint.
Right portion displays
alternating:
“HI”
“SET”
Left portion displays the
auxiliary alarm setpoint.
Right portion displays
alternating:
“AX”
“SET”
Left portion displays
the calibration gas
concentration setting .
Right portion displays
alternating:
“CAL”
“SET”
Function
Description/Action
• Range must match that of the sensor being used
• Press the INCREASE button to increase the range
setting or the DECREASE button to decrease the
setting (see Table 4 for available ranges).
Select the Range (see Table 5).
Note: Not all U9500 models support
selectable full scale range.
• When the desired setting is displayed, press the
SETUP/ACCEPT button to program the setting into
the transmitter.
• Press the INCREASE button to increase the setpoint
or the DECREASE button to decrease the setpoint.
Select the Low Alarm Setpoint
• When the desired setting is displayed, press the
SETUP/ACCEPT button to program the setpoint into
the transmitter.
• Press the INCREASE button to increase the setpoint
or the DECREASE button to decrease the setpoint.
Select the High Alarm Setpoint
• When the desired setting is displayed, press the
SETUP/ACCEPT button to program the setpoint into
the transmitter.
Select the Auxiliary Alarm Setpoint
“DE-ENERG”
or
“ENERG”
10.1
• When the desired setting is displayed, press the
SETUP/ACCEPT button to program the setpoint into
the transmitter.
• The calibration gas concentration must match that
of the gas being used to calibrate (unless K factors
for combustible sensors are being applied - refer to
“Catalytic Sensor Calibration” section.
Select
the
Concentration
Calibration
Gas
• Press the INCREASE button to increase the setting
or the DECREASE button to decrease the setting.
• When the desired setting is displayed, press the
SETUP/ACCEPT button to program the setting into
the transmitter.
Note: Not present for
Hydrocarbon (PointWatch).
“LATCH”
or
“NonLATCH”
• Press the INCREASE button to increase the setpoint
or the DECREASE button to decrease the setpoint.
• Press INCREASE or DECREASE to change the
setting.
Select Non-Latching or Latching
Relays
• When the desired setting is displayed, press the
SETUP/ACCEPT button to program the setting into
the transmitter.
• Press INCREASE or DECREASE to change the
setting.
Select De-Energized or Energized
• When the desired setting is displayed, press the
relays.
SETUP/ACCEPT button to program the setting into
the transmitter. Factory default setting is for deenergized operation.
21
95-8444
Table 10—Infiniti Setup (Continued)
Display
“YES AUTO”
“YES CAL”
or
“NO AUTO”
“NO CAL”
Function
Select Automatic
Calibration
Description/Action
or
Manual
Note: Not present for
Hydrocarbon (PointWatch)
“YES EXT”
“YES CAL”
or
“NO EXT”
“NO CAL”
• Press INCREASE or DECREASE to change the
setting.
• When the desired setting is displayed, press the
SETUP/ACCEPT button to program the setting into
the transmitter. Factory default setting is for auto
calibration.
Option to allow external reset button • Press INCREASE or DECREASE to change the
setting.
to be used for calibration in addition
to or instead of Cal Magnet. When • When the desired setting is displayed, press the
“Yes” is selected, the external reset
SETUP/ACCEPT button to program the setting into
input functions exactly like the Cal
the transmitter. Factory default setting is for no Magnet.
external reset button is not used for calibration.
• Press the SETUP/ ACCEPT button to bypass the 4
to 20 mA calibration procedure and exit the SETUP
mode.
“NO 4-20”
“NO CAL”
Option to select a Current Loop
Output other than factory calibrated • Press DECREASE or INCREASE to change the
display to read “YES 4-20 CAL.”
4 to 20 mA.
• Press the SETUP/ACCEPT button to program the
current loop.
IMPORTANT
A dc current meter capable of measuring 4 to 20 mA must be connected to the current loop output for the following
three current loop adjustments. This can be accomplished by connecting a dc ammeter in series with the load
or by connecting a digital dc voltmeter across a known load resistance and calculating the current flow using the
formula: current (I) = voltage/load resistance. Do not press INCREASE or DECREASE pushbuttons during Current
Loop Cal without a current meter or voltmeter connected to the current loop output as described above. Doing so
will result in a non-calibrated current loop. Before performing any current loop adjustment sequences, be aware
that a PLC connected to a 4-20mA output could be activated, causing an alarm.
“SET 4mA”
“CURRENT”
“SET 20mA”
“CURRENT”
“SET CAL”
“CURRENT”
Set the current output zero level.
• Press the INCREASE or DECREASE button to
change the zero level current output (indicated on
the dc current meter).
• When the desired output is indicated, press the
SETUP/ACCEPT button to program the setting into
the transmitter.
• Press the INCREASE or DECREASE button to
change the full scale current level output (indicated
Set the current output full scale
on the dc current meter).
level.
• When the desired output is indicated, press the
SETUP/ACCEPT button to program the setting into
the transmitter.
• Press the INCREASE or DECREASE button to change
the calibration current level output (indicated on the
dc
current meter).
Set the current output during
calibration and setup modes.
• When the desired output is indicated, press the
SETUP/ACCEPT button to program the setting into
the transmitter.
10.1
22
95-8444
For toxic gas sensor calibration, the calibration gas
type must match the sensor cell type. The calibration
gas concentration should be one half of the full scale
measurement range of the sensor in use.
Table 11—Recommended Sensor Calibration Methods
Calibration
Method
Sensor
Auto
Combustible Catalytic (Ex)
X
H2S
X
CO
X
PointWatch (HC)
X only
O2
X only
Manual
Cl2
X only
SO2
X only
For oxygen sensor calibration, normal ambient air may
be used for span calibration if it is known to be clean and
not depleted of O2 (21% vol.). Compressed clean air is
available from Det-Tronics if required. It is not necessary
to utilize an O2 free gas such as compressed nitrogen
to calibrate the O2 sensor zero level. An electronic zero
switch is provided within the C7065E sensor housing for
this function. Refer to the O2 sensor instruction manual
for details.
Important Calibration Notes
Calibration Gas Sources
For optimum performance,
calibration gases.
only
use
Det-Tronics
IMPORTANT
The calibration gas concentration setting must
match the concentration of the gas used to
perform the calibration (printed on label of
calibration gas cylinder).
For PointWatch calibration, use only 50% LFL
concentration calibration gas of the same type as
selected by the PointWatch gas selection switch. Refer
to the PointWatch instruction manual for details.
•
Ensure that the transmitter is properly programmed
for the gas concentration being used for calibration.
(See the “Setup” section.)
•
Ensure that only clean air is present at the sensor prior
to entering the calibration mode. If the possibility
of background gases exists, purge the sensor with
clean air to ensure accurate calibration.
•
Ensure that the concentration of the calibration
gas matches the calibration gas setting. Typically,
calibration gas concentration equal to 50% full scale
is used.
•
If the calibration procedure is not completed or
if the sensitivity of the sensor has deteriorated to
the extent that calibration cannot be successfully
completed, a fault will be generated and the system
will automatically revert back to the former calibration
settings (after 10 minutes or when the gas level drops
below the lowest setpoint). If a successful calibration
cannot be accomplished, replace the sensor and the
hydrophobic filter (if used) and re-calibrate.
For catalytic sensor calibration, most applications
require the use of 50% LFL calibration gas matching the
type of gas to be detected. If a matching calibration gas
type is not available, a conversion K-factor must be used.
Refer to the “Catalytic Sensor Calibration” section in this
manual for details.
Table 12—Recommended Frequency of Calibration for Det-Tronics Sensors
Sensor Type
After 1 Hour Warm-up
After 24 Hours
After 1 Week
Every 90 Days
Opt
Opt
Catalytic
*
Req
Opt
Req
Req
Req
H2S Electrochemical
Req
Opt
Opt
Opt
Chlorine
Req
Req
NR
Req
Oxygen
Req
Opt
Opt
Opt
PointWatch
Carbon Monoxide
Req
Opt
Opt
Opt
Sulfur Dioxide
Req
Opt
Opt
Opt
Req = Required
Opt = Optional
NR = Not Recommended
*
= Optional for methane. Must be calibrated if gas selection setting other than methane is used.
10.1
23
95-8444
•
Inspect the sensor. Loss of sensitivity can be
caused by various factors. One common cause is
by clogging of the hydrophobic or sintered filters by
dirt, oil, paint, etc. Problems of this nature will not
be detected by the transmitter’s diagnostic circuitry
when used with catalytic or electrochemical sensors.
If the filter or the sensor are dirty or plugged, they
should be replaced. When the transmitter is used
with the PointWatch Infrared detector, contamination
of its optical surfaces will be annunciated.
•
When calibrating a C7065E O2 Sensor, the initial
“zero” calibration requires activation of a switch
located inside the sensor housing. Subsequent
calibrations of the sensor do not require switch
activation. Always calibrate the sensor using 20.9%
O2 if ambient air is less than 20.9% O2.
A conversion K-factor must be used in the calibration of
the system whenever detection of gases/vapors other
than the gas used in the actual calibration process will
occur. Refer to the “Det-Tronics K-factor Tech Note” for
the current list of Det-Tronics K-factors. The K-factor
represents the relative sensor response ratio of the
calibration gas to the detected gas. The K-factor is used
within the “K-factor equation” to determine the proper
transmitter output level (span setpoint) when the sensor
is exposed to the calibration gas. The K-factor equation
is as follows:
CxK=S
AUTO CALIBRATION PROCEDURE
In the Auto Cal mode, all adjustments are made
automatically by the transmitter once calibration has
been initiated. Auto Cal mode is recommended for all
sensors except Cl2 and SO2. Refer to Table 13 for the
Automatic Calibration Procedure.
Concentration of the calibration gas in % LFL
K =
Conversion K-factor for the gas to be monitored
with the given calibration gas
S =
Corrected combustible gas transmitter
calibration span output level (span setpoint)
Example: Gas to be detected: Propane
Calibration gas:
50% LFL Methane in air
K-factor: 1.39
Equation:
O2 Sensor Calibration
Initial calibration of the U9500C/C7065E O2 detector
requires removal of the C7065E sensor housing for access
to the zero selection switch. Subsequent calibrations of
the sensor will not require removal of the sensor housing,
nor nitrogen purging. The U9500C will automatically set
the zero cal level even with normal levels of O2 displayed
during the zero calibration process. Refer to “Calibration
Procedure for Infiniti with C7065E Oxygen Sensor” for
details.
C x K = S
50 x 1.39 = 69.5
For proper system sensitivity to propane, the Infiniti
transmitter calibration gas concentration setpoint should
be adjusted to read 69% LFL. When the sensor is
calibrated with 50% LFL methane in air calibration gas
mixture, the sensor will provide an accurate measurement
of propane.
If more than one calibration gas type and K-factor is
listed for a gas that will be detected, generally the
best calibration gas to use is that which provides a
K-factor closest to the value of 1.0 (one). Note that the
presence of Methane gas as a potential detectable gas
is an exception. Always use Methane calibration gas if
Methane is expected to be present within the protected
area.
Catalytic Sensor Calibration
Catalytic sensors respond to many combustible gases.
While it is highly recommended that a sensor be calibrated
to detect a single gas, the fact remains that only a few
calibration gas types are available. In addition, multiple
gas hazards may exist.
If a specific gas/vapor and K-factor is not listed within
the “Det-Tronics K-factor Tech Note”, a K-factor can
be determined by Det-Tronics Sensor Engineering
department. Contact the factory for details.
A known concentration (typically one half of the full scale
detection range of the instrument) of the actual gas or
vapor type that is expected to be detected should be
used to calibrate the system. Det-Tronics calibration
gas should always be used to ensure proper system
calibration performance and accuracy. Calibration gas
should not be used if the Oxygen concentration within
the gas is listed at less than 20% by volume.
10.1
C =
24
95-8444
Table 13—Automatic Calibration Procedure
Description
Display
Operator Action
Normal operation/no gas present
Indicates the detected gas
concentration
• If the possibility of background gases exists,
purge the sensor with clean air to ensure accurate
calibration.
Initiate calibration – Catalytic,
PIR9400 & all toxic sensors
New sensor cal – O2
Existing sensor cal – O2
Zero calibration complete
New sensor cal – O2
Span calibration
Sequences through the Setup
settings
Enters the Calibration mode
- Display shows detected gas
concentration and alternating
message:
“ZERO” “CAL”
• Hold the Cal magnet to the CAL/RESET magnetic
switch for 7 seconds. See Figure 18 for CAL/RESET
switch location. (The external reset input can also be
used to enter the calibration mode if “YES EXT CAL”
was selected during Setup procedure).
New O2 sensor cal – First remove C7065E cover and
select “Zero”.
No zero gas action necessary.
Display shows detected gas
concentration and alternating
message:
“APLY”
“GAS”
• Apply the calibration gas to the sensor by placing
the calibration cup over the sensor (or installing
the direct inject fitting and hose on the PointWatch
detector) and opening the valve on the calibration
gas cylinder.
Select switch to “ON”. Replace C7065E cover.
Display
shows
rising
gas
concentration
and
alternating
• None.
message:
“GAS”
“ON”
Span calibration complete
• Remove the calibration gas. When the gas level
falls below the lowest alarm setpoint, and if no faults
are present, the transmitter automatically exits the
Calibrate mode.
Display shows decreasing gas
concentration when gas is removed
• A sensitivity reading that can be used to track
and alternating message:
sensor life (except PointWatch) is displayed for
“CAL” “OK”
seven seconds with the “SPAN” reading and prior
then
to the transmitter returning to normal operation. Any
“RMV”
reading over 100 indicates that the sensor is good.
“GAS”
then, if the calibration is successful:
• At the successful completion of the calibration, all
“XXXX SPAN”
outputs and indicators return to normal operation.
• If faults are present, the unit will exit after the
remainder of 10 minutes.
Calibration fault indication
10.1
Display shows alternating messages • If a fault occurs, remove the gas and correct the fault.
indicating the fault, then
After the fault has been cleared, begin calibration
again.
See Table 15 - Fault Messages, Explanations
“RMV”
and
Corrective
Action.
“GAS”
25
95-8444
5. For best results, allow the U9500/C7065E to warm
up for one hour before performing the calibration
procedure.
MANUAL CALIBRATION PROCEDURE
In the Manual Cal mode, the CAL/RESET switch must be
activated using the Cal magnet to accept a reading as a
calibration point. (The external reset input can also be
used to accept a reading if “YES EXT CAL” was selected
during the Setup procedure.)
Initial/Replacement Sensor Calibration Procedure
1. Remove the C7065E aluminum sensor cover. Leave
the sensor cell plugged in.
The Infiniti must be programmed for manual calibration
prior to initiating the manual calibration procedure (refer
to the Setup section). Refer to Table 14 for the Manual
Calibration Procedure.
2. Apply the calibration magnet to the U9500 calibration
target location for nine seconds, or until the U9500
display shows “zero-cal”.
3. Move the toggle switch on the electrochemical
sensor cell to the “zero” position. See Figure 19 for
switch location. This will simulate the “zero” or “nooxygen” condition necessary to establish a baseline
output calibration level. When the U9500 display
reads “00”, move the toggle switch back to the “on”
position. Place the aluminum sensor cover over the
cell and reinstall.
IMPORTANT
The Manual Calibration Procedure must be used
for chlorine (Cl2) and sulfur dioxide (SO2) sensors.
Use the Auto Calibration Procedure for all other
sensors.
Calibration Procedure for Infiniti with
C7065E Oxygen Sensor
There are two calibration procedures for the U9500/
C7065E:
1. Initial/Replacement Sensor Calibration.
This
procedure should be used on new product initial
calibration, or any time an existing sensor has
been replaced with a new or different sensor. This
procedure requires removal of the sensor housing.
Live DC circuits will be exposed to the surrounding
environment during this procedure.
ZERO SWITCH
2. Routine Calibration. This procedure should be
used for subsequent calibrations after the initial/
replacement sensor calibration procedure has been
performed. This procedure does not require removal
of the sensor housing, and no live circuits will be
exposed.
A1472
Figure 19—Electrochemical O2 Sensor Cell
4. When the U9500 display shows “aply-gas”, the
transmitter will begin to record the sensor output level
in response to the ambient oxygen concentration.
If this is known to be 20.9%, an accurate span
calculation will result. If the ambient O2 level is
less than 20.9%, it is recommended to apply 20.9%
oxygen using a Det-Tronics O2 calibration kit.
Before performing either procedure:
1. Ensure that the device is properly installed and
wired.
2. Ensure that the electrochemical cell is installed
properly.
3. Ensure that the U9500/C7065E is powered-up
and displays normal operating mode without fault
indication. In not, clear all faults before attempting
to enter calibration mode.
5. When span calculations are complete, the U9500
display will indicate “cal-OK”, and then “rmv-gas”.
If using a Det-Tronics O2 calibration kit, remove the
calibration cup and close the regulator valve. No
actions are necessary if ambient air O2 is used.
4. Ensure that the U9500/C7065E has been
programmed for “auto calibration” mode. Manual
calibration mode is not compatible with the C7065E
sensor.
6. The U9500 will automatically exit calibration mode
when calibration is complete. The displayed O2 level
should be 20.9% (17.4 mA current level) if ambient
conditions are normal.
10.1
26
95-8444
5. While the Infiniti transmitter display is indicating “Zero
Drift,” simultaneously press and release the Increase
and Decrease switches on the Infiniti faceplate. The
Infiniti will exit the warm up mode and display an
incorrect gas level reading.
Routine Calibration Procedure
1. Do not remove the C7065E sensor cover.
2. Apply the calibration magnet to the U9500 calibration
target location for nine seconds, or until the U9500
display shows “zero-cal”. The initial zero baseline
output level reading will be referenced. Wait until the
U9500 display shows “aply-gas”.
6. Calibrate the U9500A following the “Initial/
Replacement Sensor Calibration” procedure.
7. Replace the Infiniti J-box cover.
3. The transmitter will now begin to record the sensor
output level in response to the ambient oxygen
concentration. If this is known to be 20.9%, an
accurate span calculation will result. If the ambient
O2 level is less than 20.9%, it is recommended
to apply 20.9% oxygen using a Det-Tronics O2
calibration kit.
8. Re-enable any alarming devices triggered by the
U9500A Infiniti alarm outputs.
Alternate method:
This method can be used in a classified/hazardous area.
1. Disable any alarming devices triggered by the
U9500A Infiniti transmitter alarm outputs.
4. When span calculations are complete, the U9500
display will indicate “cal-OK”, and then “rmv-gas”.
If using a Det-Tronics O2 calibration kit, remove the
calibration cup and close the regulator valve. No
actions are necessary if ambient air O2 is used.
2. Power up the U9500A with a new or different CGS
sensor installed.
3. Wait for the U9500A Infiniti transmitter to display
“Zero Drift” indication (approximately 4 minutes
from power up).
5. The U9500 will automatically exit calibration mode
when calibration is complete. The displayed O2 level
should be 20.9% (17.4 mA current level) if ambient
conditions are normal.
4. Apply calibration gas to the sensor. After
approximately 30 seconds, the U9500A will reset,
go through the startup cycle, and display some
incorrect gas level.
cgs sensor replacement
Under certain circumstances, an existing U9500A
Infiniti gas transmitter fitted with a replacement CGS
sensor may not exit “warm up” mode after four minutes
from power up, causing a “Zero Drift” indication to be
displayed on the U9500A. This uncommon situation can
occur when a CGS pellistor resistance is slightly offset
from the previous CGS pellistor resistance. It does not
mean that the sensor or the Infiniti transmitter are bad or
out of tolerance.
5. Apply a calibration magnet and hold it in place until
the Infiniti transmitter enters the calibration mode.
The Infiniti display will read a gas indication number
alternating with “Zero Cal”.
6. Remove the calibration gas from the sensor.
7. Allow the Infiniti to zero calibrate to the new or
different sensor.
8. When prompted by the Infiniti, apply the calibration
gas to set the 50% LFL span setting.
There are two simple methods to solve this problem.
9. Once the Infiniti accepts the new span setting, follow
the prompts to exit the calibration mode.
Preferred method:
10. From this point, allow the sensor to fully warm up for
one hour minimum, and then calibrate the U9500A
following the normal calibration procedure in the
instruction manual.
This method requires a Hot Work Permit and
declassification of the work area.
1. Disable any alarming devices triggered by the
U9500A Infiniti transmitter alarm outputs.
11. Re-enable any alarming devices triggered by the
U9500A Infiniti alarm outputs.
2. Power up the U9500A with a new or different CGS
sensor installed.
Contact the factory if you have any questions or
comments.
3. Wait for the U9500A Infiniti transmitter to display
“Zero Drift” indication (approximately 4 minutes
from power up).
4. Remove the cover from the Infiniti transmitter.
10.1
27
95-8444
Table 14—Manual Calibration Procedure
Description
Normal operation/no gas present
Display
Operator Action
• Ensure that the unit is set for Manual Calibration.
Refer to the “Setup Procedure.”
Indicates the detected gas
concentration
• If the possibility of background gases exists,
purge the sensor with clean air to ensure accurate
calibration.
Sequences through the Setup
settings
Initiate calibration
Enters the Calibration mode
- Display shows detected gas
concentration and alternating
message:
• Hold the Cal magnet to the CAL/RESET magnetic
switch for 7 seconds to enter the calibration mode.
(The external reset input can also be used to enter
the calibration mode if “EXT CAL YES” was selected
during the Setup procedure).
“ZERO”
“CAL”
Zero calibration
Apply calibration gas
Display continues showing detected • When the zero reading has stabilized, hold the Cal
magnet to the CAL/RESET magnetic switch for 2
gas concentration and alternating
seconds to accept the displayed reading as the zero
message:
point.
“ZERO”
“CAL”
Display shows detected gas
concentration and alternating
message:
“APLY”
“GAS”
Span calibration
Display shows rising gas
concentration and alternating
message:
“GAS”
“ON”
• Apply the calibration gas to the sensor by placing
the calibration cup over the sensor and opening the
valve on the calibration gas cylinder.
• When the reading has stabilized, hold the Cal magnet
to the CAL/RESET magnetic switch for 2 seconds
to accept the displayed reading as the span point.
The display will indicate the selected calibration gas
concentration as span.
• Remove the calibration gas. When the gas level
falls below the lowest alarm setpoint, the transmitter
automatically exits the Calibrate mode.
Remove calibration gas
Calibration fault indication
10.1
Display shows decreasing gas • A sensitivity reading that can be used to track sensor
life (except PointWatch) is displayed for 7 seconds
concentration when gas is removed
with the “SPAN” display and prior to the transmitter
and alternating message:
returning to normal operation. Any reading over 100
“CAL” “OK”
indicates that the sensor is good.
then
“RMV”
• At the successful completion of the calibration, all
“GAS”
outputs and indicators return to normal operation.
then
• If faults are present, the unit will exit after the
“XXXX SPAN”
remainder of 10 minutes. To force an exit from
calibration, make sure that the gas concentration is
below alarm setpoints and dropping, then activate
the CAL/RESET switch for 1 second.
Display shows alternating messages • If a fault occurs, remove the gas and correct the fault.
indicating the fault, then
After the fault has been cleared, begin calibration
again. See Table 15 - Fault Messages, Explanations
“RMV”
and Corrective Action.
“GAS”
28
95-8444
TROUBLESHOOTING
Prioritized Faults
The faults are prioritized, with the highest fault being the
only one displayed (see Table 15 for a prioritized listing).
If an additional fault exists, it will be displayed after the
higher priority fault has been cleared.
Automatic Diagnostics and Fault Messages
The transmitter features self-testing circuitry that
continuously checks for problems that could prevent
proper system response. When power is applied, the
microprocessor automatically tests the unit. If a fault is
detected during this warmup mode, it will be indicated
on the display and must be reset to clear. In the Normal
operating mode, it continuously monitors the input signal
from the sensor to ensure proper functioning.
Clearing Faults
Generally, faults that occur while in the normal mode are
self-clearing once the fault condition has been corrected.
Faults that occur while in the warmup and calibration
modes require a manual reset to clear (using either an
externally connected reset switch or the Cal magnet).
After the fault condition has been corrected, the fault
relay automatically switches to the normal (energized)
state, the dc current output returns to normal, and the
fault message turns off.
In the event of a system fault:
•
The display identifies the nature of the fault using a
fault message. Refer to Table 15 for an expanded
explanation of the messages.
•
The normally energized Fault relay (on relay models)
is de-energized.
•
The dc current output drops to less than 1.0 mA.
CAUTION
The fault detection circuitry does not monitor the
operation of external response equipment or the
wiring to these devices. It is important that these
devices be checked periodically to ensure that
they are operational.
NOTE
The fault message will be shown for about
1.5 seconds out of every 3 seconds. The gas
concentration at the sensor will be displayed
during the remaining time. If more than one fault
should occur, the highest priority fault will be
displayed. (Table 15 lists the faults in order of
priority.)
In the event of an alarm condition and a system
fault both occurring:
•
In most cases, the first condition that occurs will be
indicated by the current and relay outputs and on
the display.
•
The exceptions are “CAL ABORTED” and “SENSOR
E.O.L.” faults, which both can occur during the
calibration procedure. If an alarm occurs with these
faults, the alarm will over-ride the fault and will be
indicated.
10.1
29
95-8444
Table 15—Fault Messages, Explanations And Corrective Action
Fault Message Display
Blank Display
EEPROM sumcheck failure. Cycle power to clear fault. If fault persists, return the unit to the
factory for repair.
“COMPUTER”
“FAILURE”
RAM or processor failure. Cycle power to clear fault. If fault persists, return the unit to the factory
for repair.
“WDT”
“FAILURE”
Watchdog timer failure. Cycle power to clear fault. If fault persists, return the unit to the factory
for repair.
“EEPROM”
“FAILURE”
Activate the Cal/Reset switch using the Cal Magnet, then perform Setup and Calibration
procedures. If fault persists, return the unit to the factory for repair.
“EXT RSET”
“PROBLEM”
External reset button has been activated for 15 seconds or longer. Self-clearing when button is
released.
“24V P.S.”
“FAILURE”
External 24 volt power supply is not in the 15 to 32 volt range. Check and correct input voltage.
During normal operating mode, this fault is self clearing when the fault is corrected. If fault
occurs during warmup or calibration modes, activate the Cal/Reset switch using the Cal Magnet.
“5V P.S”
“FAILURE”
Internal 5 volt analog power supply is not in the 4.75 to 5.35 volt range. During normal operating
mode, this fault is self clearing when the fault is corrected. If fault occurs during warmup or
calibration modes, activate the Cal/Reset switch using the Cal Magnet. If fault persists, return
the unit to the factory for repair.
“SENSOR”
“PROBLEM”
Sensor input fault. With 4 to 20 mA output sensors, the current output is below 2 mA or above
35 mA. In normal operating mode, the unit automatically goes through warmup when this fault
clears. If this fault occurs at the end of the warmup period or calibration procedure, recalibrate
the sensor. If fault persists, check sensor condition and wiring.
“PTIR CAL”
“LINE LOW”
(PointWatch sensor only). The PointWatch has detected that the calibration line is low (active)
during power-up.
“PTIR”
“VOLT LOW”
(PointWatch sensor only). The PointWatch has detected that its input voltage is below specified
level.
“DIRTY”
“OPTICS”
(PointWatch sensor only). The PointWatch has detected that its optics are dirty. Refer to the
PointWatch manual for cleaning procedure.
“REPLACE”
“SENSOR”
(In calibration mode) sensor is defective. Replace sensor and perform calibration procedure.
“CAL”
“ABORTED”
(Cal Message) Time ran out while waiting for the gas reading to stabilize. Activate the Cal/Reset
switch using the Cal Magnet.
“SENSOR”
“E.O.L.”
(Cal Message) Sensor reaching End Of Life. Consider replacement of the sensor in the next 1
or 2 calibrations.
“ZERO”
“DRIFT”
10.1
Explanation and Corrective Action
Negative zero drift. Sensor input is –9% full scale or lower. Perform sensor calibration.
30
95-8444
Contaminating Gases And Vapors for Catalytic
Sensors
•
Fault detection circuitry continuously monitors for
problems that could prevent proper system response.
It does not monitor external response equipment or
the wiring to these devices. It is important that these
devices be checked initially when the system is
installed, as well as periodically during the ongoing
maintenance program.
•
The system must be checked periodically in the
Normal mode to ensure that those items not checked
by the transmitter diagnostic circuitry (such as output
relays) are functioning properly.
Gases and/or vapors that can cause permanent loss of
sensor sensitivity:
–– Silicone compounds often found in oils, greases
and resins.
–– Antiknock compounds such as tetra ethyl lead,
tetra methyl lead and phosphate esters.
Gases and/or vapors that can cause inhibition (temporary
loss of sensitivity) of the sensing element:
SENSOR AND HYDROPHOBIC FILTER
REPLACEMENT
–– Volatile halogenated compounds.
–– Tetrachloroethylene.
CAUTION
Be sure to secure all output devices that are
actuated by the system to prevent unwanted
activation of this equipment, and remember to
place these same output devices back into service
when the checkout is complete.
–– Fluorinated hydrocarbon.
–– Vapors that result from hot P.V.C.
–– Acids and various solvents may cause a shift in the
zero of electrochemical sensor.
–– Chlorine and Chlorinated HC.
Catalytic Sensor Replacement
NOTE
The sensor must be periodically inspected and
calibrated following prolonged exposure to
contaminating gases and vapors.
For catalytic combustible gas sensors, it is necessary to
replace the entire sensor assembly when it is defective.
Follow the procedure below to replace the sensor.
1. Remove power to the transmitter prior to replacing
the sensor.
MAINTENANCE
2. Remove the transmitter cover and the wire shield
within the transmitter. If a sensor termination box
was used, remove the sensor junction box cover.
CAUTION
Be sure to secure all output devices that are
actuated by the system to prevent unwanted
activation of this equipment, and remember to
place these same output devices back into service
when the checkout is complete.
•
•
10.1
3. Unplug the sensor from the transmitter module (if a
sensor termination box was used, unplug it from the
plug inside the termination box) and unscrew it from
the conduit entry.
4. Thread the wires for the replacement sensor through
the conduit entry, then screw the sensor into the
conduit entry and plug it in. Replace the wire
shield.
Periodically apply a generous amount of Lubriplate
grease (part number 102868-001) to the internal
threads on the junction box cover using an
application tool. DO NOT use other lubricants, since
some materials can cause irreversible damage to the
sensing element in the gas sensor.
5. Replace the junction box cover.
6. Re-apply power. Allow time for the unit to warm
up and stabilize (approximately 24 hours for best
results), then calibrate.
To ensure reliable protection, it is important to check
and calibrate the detection system on a regularly
scheduled basis. The frequency of these checks
is determined by the requirements of the particular
installation - 90 days is typical.
31
95-8444
Electrochemical Sensor Cell Replacement
NOTE
Handle the sensor cell carefully. To avoid
possible damage, observe the normally accepted
procedures for handling electrostatic sensitive
devices. See form 75-1005 for additional
information.
For sensors with replaceable sensor cells (toxic and
oxygen sensors), it may not be necessary to replace
the entire sensor housing assembly when the sensor is
defective. Follow the procedure below to replace the
sensor cell.
1. Remove power to the transmitter prior to replacing
the sensor cell.
5. Be sure that the O-ring on the sensor housing is
in good condition, then place the cap back on the
sensor base. Tighten only until snug. Do not over
tighten.
2. Remove the cap from the sensor base. See Figure
20. (There is no need to remove the sensor housing
from the junction box for this operation).
6. Re-apply power. Allow time for the unit to warm
up and stabilize (approximately one hour for best
results), then calibrate.
NOTE
Compare part numbers to be sure that the correct
replacement cell is being used.
An adequate supply of spare electrochemical cell
assemblies should be kept on hand for field replacement.
For maximum protection against contamination and
deterioration, they should not be removed from the
original protective packaging until the time of installation.
To ensure maximum storage life, electrochemical cells
should be stored in the unopened bag that the sensor
is shipped in, at a temperature between 32°F and 68°F
(0 to 20°C) and a relative humidity between 15 and 90
percent.
3. Remove the old sensor cell. The sensor cell contains
a small amount of lead – dispose of properly. Check
for corrosion or contamination on the terminals of the
sensor base, and clean if necessary.
4. Determine proper orientation for the new cell, then
carefully plug it in.
SENSOR HOUSING
SENSING ELEMENT ASSEMBLY
CAP
HYDROPHOBIC FILTER
NOTE: SENSOR APPEARANCE MAY VARY
SLIGHTLY DEPENDING UPON MODEL.
B1203
Figure 20—Electrochemical Sensor Parts Identification
10.1
32
95-8444
DEVICE REPAIR AND RETURN
Hydrophobic Filter Replacement
The hydrophobic filter on the front of some of the
electrochemical sensor housings protects the sensor cell
from contaminants in the environment, and also enables
the operation of the cell in “wet” environments without
plugging its screen. The operator should frequently
inspect the hydrophobic filter for cleanliness. A dirty filter
can significantly reduce the amount of gas that is able to
reach the sensor cell, thereby impairing the ability of the
system to respond to a hazardous condition. If the filter
becomes dirty or if it is damaged, it must be replaced.
Prior to returning devices, contact the nearest local
Detector Electronics office so that a Service Order
number can be assigned. A written statement describing
the malfunction must accompany the returned device to
expedite finding the cause of the failure.
Pack the unit properly. Use sufficient packing material
in addition to an antistatic bag or aluminum-backed
cardboard as protection from electrostatic discharge.
Return all equipment transportation prepaid to the factory
in Minneapolis.
To replace the hydrophobic filter, simply unscrew the
existing filter from the housing, then replace it with a new
filter. Use care not to over tighten.
NOTE
A dirty hydrophobic filter can adversely affect the
response of the sensor by blocking the flow of
gas to the sensor cell. If the detector cannot be
calibrated or responds slowly to the calibration
gas, check the condition of the hydrophobic filter
before replacing the sensor cell. The hydrophobic
filter should be clean and squarely seated in the
housing.
ORDERING INFORMATION
When ordering, please specifiy:
Model U9500 Infiniti Gas Transmitter
Refer to the U9500 Model Matrix for details
U9500 MODEL Matrix
MODEL
PointWatch Detector Replacement
U9500
Refer to Figure 9 in this manual for wiring diagram and
the PointWatch manual for installation and replacement
instructions.
DESCRIPTION
Infiniti Gas Transmitter
Type
GAS
A10
Catalytic Combustible
B20
H2S – Hydrogen Sulfide
SPARE PARTS
C30
02 – Oxygen
•
D40
CL2 – Chlorine
E50
CO – Carbon Monoxide
F60
SO2 – Sulfer Dioxide
G60
NO2 – Nitrogen Dioxide
H60
IR Combustible
Electronic Module - Specify gas sensor type and with
or without optional relay package when ordering.
Catalytic Combustible Gas
PointWatch Hydrocarbon Gas
Hydrogen Sulfide – Electrochemical
Carbon Monoxide (specify range)
Oxygen
Chlorine
Sulfur Dioxide
•
Calibration Gas Kits and spare cylinders - Specify
gas and concentration when ordering.
•
Sensor Termination Box.
•
Lubriplate grease for threads.
•
There are many other accessories available for
use with various sensors. Contact your local
representative or the factory for information.
10.1
Type
Output
01 or 02
4-20 mA
04
33
Relay and 4-20 mA
95-8444
APPENDIX A
FM APPROVAL AND PERFORMANCE REPORT
Refer to drawing 006722-001 (Figure A1) for applicable area classification and installation details.
COMBUSTIBLE
Transmitters: U9500A1001, U9500A1004, with Junction Box P/N 006264-XXX.
•
Explosion-proof for Class I, Division 1, Groups B, C and D Hazardous (Classified) Locations per FM 3615.
•
Dust ignition-proof for Class II/III, Division 1, Groups E, F and G Hazardous (Classified) Locations per FM 3615.
•
Non-incendive for Class I, Division 2, Groups A, B, C and D (T4A); Class II/III, Division 2, Groups F and G (T4A)
Hazardous (Classified) Locations per FM 3611.
•
Enclosure Rating NEMA/Type 4X (Indoor/Outdoor) per ANSI/NEMA 250. (For use with Sensor Termination box.)
•
Performance verified up to 100% LFL methane-in-air atmospheres per FM 6320.
Hydrocarbon Gas Transmitters: U9500H6002 or U9500H6004, with Junction Box P/N 006264-XXX.
•
Explosion-proof for Class I, Division 1, Groups B, C and D Hazardous (Classified) Locations per FM 3615.
•
Dust ignition-proof for Class II/III, Division 1, Groups E, F and G Hazardous (Classified) Locations per FM 3615.
•
Non-incendive for Class I, Division 2, Groups A, B, C and D (T4A); Class II/III, Division 2, Groups F and G (T4A)
Hazardous (Classified) Locations per FM 3611.
•
Enclosure Rating NEMA/Type 4X (Indoor/Outdoor) per ANSI/NEMA 250. (For use with Sensor Termination box.)
•
Performance verified up to 100% LFL methane-in-air atmospheres per FM 6320.
Sensor Termination Box: Model STB Series, used with Combustible units above:
•
Explosion-proof for Class I, Division 1, Groups B, C and D Hazardous (Classified) Locations per FM 3615.
•
Dust ignition-proof for Class II/III, Division 1, Groups E, F and G Hazardous (Classified) Locations per FM 3615.
•
Non-incendive for Class I, Division 2, Groups A, B, C and D (T4A); Class II/III, Division 2, Groups F and G (T4A)
Hazardous (Classified) Locations per FM 3611.
•
Enclosure Rating NEMA/Type 4X (Indoor/Outdoor) per ANSI/NEMA 250.
Combustible Gas Sensors: Model CGS:
•
Explosion-proof for Class I, Division 1, Groups B, C and D Hazardous (Classified) Locations per FM 3615.
Hydrocarbon Sensors:
Model PIR9400:
•
Explosion-proof for Class I, Division 1, Groups B, C and D (T5) Hazardous (Classified) Locations.
•
Class I, Division 2, Groups A, B, C and D (T3C) Hazardous (Classified) Locations, Tamb = –40°C to +75°C, conduit
seal not required.
Model PIRECL:
•
Explosion-proof for Class I, Division 1, Groups B, C and D (T4) Ta = 75°C with intrinsically safe output for HART
Communications in accordance with control drawing 0072783-001, conduit seal not required.
•
Non-incendive for Class I, Division 2, Groups A, B, C and D (T4) Ta = 75°C Hazardous (Classified) Locations.
Stand-alone 4-20 mA FM Approved Gas Sensor/Transmitter:
•
10.1
See specific detector rating.
A-1
95-8444
ELECTROCHEMICAL
Transmitters:
U9500B2002 and U9500B2004 (Hydrogen Sulfide), U9500C (Oxygen), U9500D (Chlorine),
U9500E (Carbon Monoxide) and U9500F (Sulfur Dioxide) with junction boxes P/N 006264-XXX and direct mounted
electrochemical sensor:
•
Explosion-proof for Class I, Division 1, Groups B, C and D Hazardous (Classified) Locations per FM 3615.
•
Dust ignition-proof for Class II/III, Division 1, Groups E, F and G Hazardous (Classified) Locations per FM 3615.
•
Non-incendive for Class I, Division 2, Groups A, B, C and D (T4A); Class II/III, Division 2, Groups F and G (T4A)
Hazardous (Classified) Locations per FM 3611.
•
Enclosure Rating NEMA/Type 4X (Indoor/Outdoor) per ANSI/NEMA 250. (For use with Sensor Termination box.)
•
Performance verified up to 100 ppm H2S-in-air atmospheres per FM 6341 (Draft).
Sensor Termination Box: STB Series, used with Electrochemical units above:
•
Explosion-proof for Class I, Division 1, Groups B, C and D Hazardous (Classified) Locations per FM 3615.
•
Dust ignition-proof for Class II/III, Division 1, Groups E, F and G Hazardous (Classified) Locations per FM 3615.
•
Non-incendive for Class I, Division 2, Groups A, B, C and D (T4A); Class II/III, Division 2, Groups F and G (T4A)
Hazardous (Classified) Locations per FM 3611.
•
Enclosure Rating NEMA/Type 4X (Indoor/Outdoor) per ANSI/NEMA 250.
Sensors:
C7064E5014 (H2S):
•
Explosion-proof for Class I, Division 1, Groups B, C and D Hazardous (Classified) Locations per FM 3615.
Sensors C7064E4012 or C7064E5012 (H2S):
•
Explosion-proof for Class I, Division 1, Groups C and D Hazardous (Classified) Locations per FM 3615.
Stand-alone 4-20 mA FM Approved Gas Detector:
•
10.1
See specific detector rating.
A-2
95-8444
U9500A SERIES SENSOR CONFIGURATIONS
•
Model CGS Series Combustible Gas Sensors
NOTE
Sensor cross sensitivity has not been verified by FM.
NOTE
The high temperature characteristic has not been FM verified above +75°C.
NOTE
Detector Electronics combustible gas detection K-Factors are not FM verified.
U9500B THROUGH U9500H INFINITI SERIES SENSOR CONFIGURATIONS
•
Models C7064E4012 and C7064E5012 Hydrogen Sulfide (H2S) Sensors (U9500B)
•
Model C7065E Oxygen (U9500C)
•
Model C7067E Chlorine (U9500D)
•
Model C7066E Carbon Monoxide (U9500E)
•
Model C7068E Sulfur Dioxide (U9500F)
•
Model PIR9400 or PIRECL Hydrocarbon (U9500H)
NOTE
FM Approval of the 4 to 20 mA input does not include or imply Approval of the gas detection apparatus such as
sensors, transmitters, or devices connected to the system. In order to maintain FM Approval of the system, all 4
to 20 mA gas detection instruments connected to the input must also be FM Approved.
NOTE
Metric straight thread types are for use in non-North American applications.
CALIBRATION
•
Calibration of the above listed sensors has been FM verified using the respective U9500 Transmitter with the
Det-Tronics 225130-001 (50% Methane) and 227115-001 H2S Calibration Kits.
•
The U9500B through U9500H Series Transmitters can be used with any FM Approved 4 to 20 mA device.
10.1
A-3
95-8444
A
B
C
AP
OL
IS
N
, M
USA
5 2
. 2 N O. 1
C2 2
U ST I B E
C O MB
FOR
38
54
, 5
%LFL
SETUP
ACCEPT
O
T
SERIAL NO.
DECREASE
ALARM
HI AUX LO
1 3 - 0 1 REV . G
DATE CODE
CAL/RESET
0 067
MODEL NO.
INCREASE
EX
XXXXX
CL I , D
IV 1
,
CL GRP S
I
B
,
CL , D
IV C
I
( Ta I , D 2 &
D
mb I V ,
G
=
C ON
D U N EM - 4 2 , R P C L
IT
A 0∞ G S
SE / T C RP
A
YP T
L
N E O
DET
TRONICS
MODEL U9500
E
R
PE M)
L
(F
)
AL 0 1
SA
0
(C
ST
I N 22- 01
0
67 088
06
N
IN
M
VDC
16-32
7.7 MAX
WATTS
(006630-001)
INFINITI GAS TRANSMITTER
W/ 006264-XXX
JUNCTION BOX
MODEL U9500 SERIES
T.
7
6
S
th
EA
P
OL
IS
N
, M
USA
5 2
. 2 N O. 1
C2 2
ST I B E
C O MB U
FOR
38
54
, 5
%LFL
EX
ACCEPT
SETUP
O
T
SERIAL NO.
DECREASE
ALARM
HI AUX LO
1 3 - 0 1 REV . G
DATE CODE
CAL/RESET
0 067
MODEL NO.
INCREASE
XXXXX
CL I , D
IV 1
,
CL GRP S
I
B
,
CL , D
IV C
I
( Ta I , D 2 &
D
mb I V ,
G
=
C ON
D U N EM - 4 2 , R P C L
IT
A 0∞ G S
SE / T C RP
A
YP T
L
N E O
DET
TRONICS
MODEL U9500
R
PE M)
L
(F
)
AL 0 1
SA
0
(C
ST
I N 22- 01
0
67 088
06
N
IN
M
VDC
16-32
7.7 MAX
WATTS
(006723-001)
MODEL STB SERIES (OPTIONAL)
SENSOR TERMINATION BOX
INFINITI GAS TRANSMITTER
W/ 006264-XXX
JUNCTION BOX
(006630-001)
MODEL U9500 SERIES
T.
8
7
1. FM APPROVED PRODUCT - NO MODIFICATIONS PERMITTED WITHOUT REFERENCE TO FM.
6
5
5
OR FM APPROVED 4-20mA DEVICE
MODEL PIR9400 SERIES
HYDROCARBON
INFRARED GAS SENSOR
MODEL CGS SERIES
COMBUSTIBLE GAS SENSOR
(006888-001)
4
2
2
MODEL C7064E5014
HYDROGEN SULFIDE SENSOR
(006705-001)
OR FMA APPROVED 4-20mA DEVICE
MODEL PIR9400 SERIES
HYDROCARBON
INFRARED GAS SENSOR
MODEL CGS SERIES
COMBUSTIBLE GAS SENSOR
(006888-001)
MODEL C7064E5014
HYDROGEN SULFIDE SENSOR
(006705-001)
CLASS I, DIVISION 1, GROUPS B,C & D
CLASS I, DIVISION 2, GROUPS A,B,C & D (T4A)
HAZARDOUS LOCATIONS
4
Figure A1—Device Configuration for FM Approval
2. WARNING: ENSURE SENSOR HAZARDOUS (CLASSIFIED) LOCATION RATING IS APPLICABLE FOR THE INTENDED USE.
FROM FM APPROVED
CONTROL DEVICE
FROM FM APPROVED
CONTROL DEVICE
S
th
00
0
11
O
.
W
1
. 6
90
RP
CO
CS
ON I
D
00
0
11
O
.
W
1
. 6
90
RP
CO
CLASS I, DIVISION 1, GROUPS B,C & D
CLASS I, DIVISION 2, GROUPS A,B,C & D (T4A)
CLASS II, DIVISION 1 GROUPS E,F & G
CLASS II, DIVISION 2, GROUPS F & G (T4A)
CLASS III, DIVISIONS 1 & 2 HAZARDOUS LOCATIONS
NEMA/TYPE 4X
E
N
IO G
N
8
A
N
A-4
WA RN I
, C L I II
NG
& G
EXP L O S : D O N
IV
E,F
M A Y BE E G AS O T O
PS
P
P RE
A
ED
GR 4 A )
AND
S E N TMO EN
APPROV
(T )
MA NC E
REA D
1,
F OR
SP W
T.
&
A
P E R S L OCA T I ON S
I V & D( T 4
MA N U UNDER
R
H E HE
F O A RD OU
AL
S
RE N
Z
BE TAN
, D ,C G
HA
FO D
A
RE I N
, B F & ∞C)
OP S TR
5
ED
ER U C
S +7
IR
A T
U
T
I
4X Q
R
CS
II
ON I
N
IO G
N
,
T OR EL EC T R
E
D E TE C
A
N
10.1
WA RN I
, C L I II
NG
& G
EXP L O S : D O N
IV
E,F
M A Y BE E G AS O T O
PS
P
P RE
A
ED
GR 4 A )
AND
S E N TMO EN
APPROV
(T )
MA NC E
REA D
1,
F OR
SP W
T.
&
A
P E R S L OCA T I ON S
I V & D( T 4
MAN U UNDER
H E HE
R
F O A RD OU
AL
S
RE N
Z
BE TAN
, D ,C G
HA
FO D
A
RE I N
, B F & ∞C)
OP S TR
5
ED
ER U C
S +7
IR
A T
U
T
I
4X Q
R
T OR EL EC T R
II
D E TE C
,
95-8444
3
3
OR FM APPROVED 4-20mA DEVICE
2
2
2
MODEL C7064E SERIES
HYDROGEN SULFIDE SENSOR
(006705-001)
OR FM APPROVED 4-20mA DEVICE
MODEL C7064E SERIES
HYDROGEN SULFIDE SENSOR
(006705-001)
1
CLASS I, DIVISION 1, GROUPS C & D
CLASS I, DIVISION 2, GROUPS A,B,C & D (T4A)
HAZARDOUS LOCATIONS
1
A
B
C
D
APPENDIX B
CSA APPROVAL
Refer to drawing 006880-001 (Figure B1) for applicable area classification and installation details.
COMBUSTIBLE
Transmitters: U9500A1001, U9500A1004, with Junction Box P/N 006264-XXX.
•
Explosion-proof for Class I, Division 1, Groups B, C and D Hazardous (Classified) Locations per CSA C22.2 #30,
and Ex d IIC T4 per CSA E60079-0, -1.
•
Dust ignition-proof for Class II/III, Division 1, Groups E, F and G Hazardous (Classified) Locations per CSA C22.2 #25.
•
Non-incendive for Class I, Division 2, Groups A, B, C and D (T4A); Class II/III, Division 2, Groups F and G (T4A)
Hazardous (Classified) Locations per CSA C22.2 #213.
•
Enclosure Rating NEMA/Type 4X (Indoor/Outdoor) per CSA C22.2 #94. (For use with Sensor Termination box.)
•
Performance verified up to 100% LFL methane-in-air atmospheres per CSA C22.2 #152.
Hydrocarbon Gas Transmitters: U9500H6002 or U9500H6004, with Junction Box P/N 006264-XXX.
•
Explosion-proof for Class I, Division 1, Groups B, C and D Hazardous (Classified) Locations per CSA C22.2 #30,
and Ex d IIC T4 per CSA E60079-0, -1.
•
Dust ignition-proof for Class II/III, Division 1, Groups E, F and G Hazardous (Classified) Locations per CSA C22.2 #25.
•
Non-incendive for Class I, Division 2, Groups A, B, C and D (T4A); Class II/III, Division 2, Groups F and G (T4A)
Hazardous (Classified) Locations per CSA C22.2 #213.
•
Enclosure Rating NEMA/Type 4X (Indoor/Outdoor) per CSA C22.2 #94. (For use with Sensor Termination box.)
•
Performance verified up to 100% LFL methane-in-air atmospheres per CSA C22.2 #152.
Sensor Termination Box: Model STB Series, used with Electrochemical units above.
•
Explosion-proof for Class I, Division 1, Groups B, C and D Hazardous (Classified) Locations per CSA C22.2 #30.
•
Dust ignition-proof for Class II/III, Division 1, Groups E, F and G Hazardous (Classified) Locations per CSA C22.2 #25.
•
Non-incendive for Class I, Division 2, Groups A, B, C and D (T4A); Class II/III, Division 2, Groups F and G (T4A)
Hazardous (Classified) Locations per CSA C22.2 #213.
•
Enclosure Rating NEMA/Type 4X (Indoor/Outdoor) per CSA C22.2 #94.
Combustible Gas Sensors: Model CGS:
•
Explosion-proof for Class I, Division 1, Groups B, C and D Hazardous (Classified) Locations per CSA C22.2 #30.
Hydrocarbon Sensors:
Model PIR9400:
•
Explosion-proof for Class I, Division 1, Groups B, C and D (T5) Hazardous (Classified) Locations
•
Class I, Division 2, Groups A, B, C and D (T3C) Hazardous (Classified) Locations, Tamb = –40°C to +75°C, conduit
seal not required.
Model PIRECL:
•
Explosion-proof for Class I, Division 1, Groups B, C and D (T4) Ta = 75°C with intrinsically safe output for HART
Communications in accordance with control drawing 007283-001, conduit seal not required.
•
Non-incendive for Class I, Division 2, Groups A, B, C and D (T4) Ta = 75°C Hazardous (Classified) Locations.
Stand-alone 4-20 mA CSA Approved Gas Sensor/Transmitter:
• See specific detector rating.
10.1
B-1
95-8444
ELECTROCHEMICAL
Transmitters: U9500B2002 and U9500B2004 (Hydrogen Sulfide), U9500C (Oxygen), U9500D (Chlorine),
U9500E (Carbon Monoxide) and U9500F (Sulfur Dioxide) with junction boxes P/N 006264-XXX and direct mounted
electrochemical sensor:
•
Explosion-proof for Class I, Division 1, Groups B, C and D Hazardous (Classified) Locations per CSA C22.2 #30,
and Ex d IIC T4 per CSA E60079-0, -1.
•
Dust ignition-proof for Class II/III, Division 1, Groups E, F and G Hazardous (Classified) Locations per CSA C22.2 #25.
•
Non-incendive for Class I, Division 2, Groups A, B, C and D (T4A); Class II/III, Division 2, Groups F and G (T4A)
Hazardous (Classified) Locations per CSA C22.2 #213.
•
Enclosure Rating NEMA/Type 4X (Indoor/Outdoor) per CSA C22.2 #94. (For use with Sensor Termination box.)
Sensor Termination Box: Model STB Series, used with Electrochemical units above.
•
Explosion-proof for Class I, Division 1, Groups B, C and D Hazardous (Classified) Locations per CSA C22.2 #30.
•
Dust ignition-proof for Class II/III, Division 1, Groups E, F and G Hazardous (Classified) Locations per CSA C22.2 #25.
•
Non-incendive for Class I, Division 2, Groups A, B, C and D (T4A); Class II/III, Division 2, Groups F and G (T4A)
Hazardous (Classified) Locations per CSA C22.2 #213.
•
Enclosure Rating NEMA/Type 4X (Indoor/Outdoor) per CSA C22.2 #94.
Sensors:
C7064E5014 (H2S):
•
Explosion-proof for Class I, Division 1, Groups B, C and D Hazardous (Classified) Locations per CSA C22.2 #30.
C7064E4012 or C7064E5012 (H2S):
•
Explosion-proof for Class I, Division 1, Groups C and D Hazardous (Classified) Locations per CSA C22.2 #30.
Stand-alone 4-20 mA CSA Approved Gas Detector:
•
10.1
See specific detector rating.
B-2
95-8444
U9500A SERIES SENSOR CONFIGURATIONS
•
Model CGS Series Combustible Gas Sensors
NOTE
The high temperature characteristic has NOT been CSA verified above +75°C.
NOTE
Detector Electronics combustible gas detection K-factors are not CSA verified.
U9500B THROUGH U9500H INFINITI SERIES SENSOR CONFIGURATIONS
•
Models C7064E4012 and C7064E5012 Hydrogen Sulfide (H2S) Sensors (U9500B)
•
Model C7065E Oxygen (U9500C)
•
Model C7067E Chlorine (U9500D)
•
Model C7066E Carbon Monoxide (U9500E)
•
Model C7068E Sulfur Dioxide (U9500F)
•
Model PIR9400 or PIRECL Hydrocarbon (U9500H)
NOTE
CSA Approval of the 4 to 20 mA input does not include or imply Approval of the gas detection apparatus such
as sensors, transmitters, or devices connected to the system. In order to maintain CSA Approval of the system,
all 4 to 20 mA gas detection instruments connected to the input must also be CSA Approved.
NOTE
Metric straight thread types are for use in non-North American applications.
CALIBRATION
•
Calibration of the above listed sensors has been CSA verified using the respective U9500 Transmitter with the
Det-Tronics 225130-001 (50% Methane) and 227115-001 H2S Calibration Kits.
•
The U9500B through U9500H Series Transmitters can be used with any CSA Approved 4 to 20 mA device.
10.1
B-3
95-8444
A
B
C
D
P
M
AP
S,
USA
MN
ACCEPT
VDC
AP
ACCEPT
VDC
WATTS
(006630-001)
MODEL U9500 SERIES
INFINITI GAS TRANSMITTER
W/ 006264-XXX
JUNCTION BOX
1X M20
1 3 - 0 0 1 REV . E
MFG. DATE
DECREASE
ALARM
HI AUX LO
CAL/RESET
MODEL NO.
0 067
INCREASE
SETUP
MODEL U9500 XXXXX
OL
%LFL
EX
N
O
TI
C NG
NE
IN
(006630-001)
7
4
5
MODEL NTMabc
NTMOS H2S GAS SENSOR
(009800-001)
IIC
MODEL C7064E5014
HYDROGEN SULFIDE SENSOR
(005557-001)
OR CSA APPROVED 4-20mA DEVICE
MODEL PIR9400 SERIES
HYDROCARBON
INFRARED GAS SENSOR
(006367-001)
MODEL CGS SERIES
COMBUSTIBLE GAS SENSOR
(006887-001)
3
MODEL C7064E5014
HYDROGEN SULFIDE SENSOR
(005557-001)
TYPE 225629
MOS HYDROGEN SULFIDE SENSOR
(007236-001)
MODEL NTMabc
NTMOS H2S GAS SENSOR
(009800-001)
OR CSA APPROVED 4-20mA DEVICE
MODEL PIR9400 SERIES
HYDROCARBON
INFRARED GAS SENSOR
(006367-001)
MODEL CGS SERIES
COMBUSTIBLE GAS SENSOR
(006887-001)
4
3
TYPE 225629
MOS HYDROGEN SULFIDE SENSOR
(007236-001)
T4
2 CLASS I, ZONE 1, Ex d
OR
CLASS I, DIVISION 1, GROUPS B,C & D
CLASS I, DIVISION 2, GROUPS A,B,C & D (T4A)
HAZARDOUS LOCATIONS
5
Figure B1—Device Configuration for CSA certification
1X M20
6
M
MODEL STB SERIES
SENSOR TERMINATION BOX
(006723-001)
1. CSA CERTIFIED PRODUCT - NO MODIFICATIONS PERMITTED WITHOUT REFERENCE TO CSA
8
P
USA
MN
I S,
1X M20
WATTS
6
MODEL U9500 SERIES
INFINITI GAS TRANSMITTER
W/ 006264-XXX
JUNCTION BOX
1 3 - 0 0 1 REV . E
MFG. DATE
DECREASE
ALARM
HI AUX LO
CAL/RESET
MODEL NO.
0 067
INCREASE
SETUP
MODEL U9500 XXXXX
I
OL
%LFL
EX
N
O
TI
C NG
NE
IN
2 WARNING: ENSURE TRANSMITTER & SENSOR HAZARDOUS (CLASSIFIED)
LOCATION RATING COMBINATION IS APPLICABLE FOR THE INTENDED USE.
FROM CSA APPROVED
CONTROL DEVICE
FROM CSA APPROVED
CONTROL DEVICE
T4
IIC
2 CLASS I, ZONE 1, Ex d
OR
CLASS I, DIVISION 1, GROUPS B,C & D
CLASS I, DIVISION 2, GROUPS A,B,C & D (T4A)
CLASS II, DIVISION 1 GROUPS E,F & G
CLASS II, DIVISION 2, GROUPS F & G (T4A)
CLASS III, DIVISIONS 1 & 2 HAZARDOUS LOCATIONS
NEMA/TYPE 4X
7
DET EC
TOR
EL E
CT
8
RO
DET EC
TOR
EL E
CT
B-4
NI
RE AD &
CS
MANUA UNDER
L B ST
CO
EF AN
OR D
R
E
,
OP I N S
ER TR
AT U
I
RO
NI
RE AD &
CS
MANUA UNDER
L B ST
CO
EF AN
OR D
R
E
,
OP I N S
ER TR
AT U
I
10.1
95-8444
MODELS C7065E
OR C7066E SERIES
OR CSA CERTIFIED 4-20mA DEVICE
MODELS C7065E
OR C7066E SERIES
1
MODELS C7064E, C7068E,
OR C7069E SERIES
OR CSA CERTIFIED 4-20mA DEVICE
2
IIC
1
MODELS C7064E, C7068E,
OR C7069E SERIES
T4
2 CLASS I, ZONE 1, Ex d
OR
CLASS I, DIVISION 1, GROUPS C & D
CLASS I, DIVISION 2, GROUPS A,B,C & D (T4A)
HAZARDOUS LOCATIONS
2
A
B
C
D
APPENDIX C
ATEX / CE APPROVAL
INFINITI U9500 GAS TRANSMITTER
0539 II 2 G
Ex d IIC T5-T6 Gb
DEMKO 02 ATEX 131327X
T6 (Tamb = –55°C to +60°C)
T5 (Tamb = –55°C to +75°C)
IP66
FM
®
APPROVED
EN Standards:
EN 50270: 2006
EN 60079-0: 2009
EN 60079-1: 2007
EN 60529: 1991+A1: 2000
EN 60079-29-1: 2007
Combustible gas performance certified per EN 60079-29-1: 2007 when used with Model CGS or any 4-20 mA stand
alone combustible gas detector.
Read and understand instruction manual before operating.
To obviate the risk of hot spots and capacitor energy storage, the enclosure must not be opened, even when isolated,
when an explosive gas atmosphere is present.
All cable entry devices and blanking elements shall be certified in type of explosion protection flameproof enclosure
“d”, suitable for the conditions of use and correctly installed. Unused apertures shall be closed with suitable certified
blanking elements.
For ambient temperatures below –10°C and above +60°C use field wiring suitable for both minimum and maximum
ambient temperature.
ATEX Special Conditions for Safe Use (X):
The Infiniti Transmitter types U9500A and U9500H have an ambient temperature rating for performance of –40°C to +75°C.
The measuring function of the Infiniti Transmitter types U9500B, U9500C, U9500D, U9500E, U9500F and U9500G for
explosion protection, according to Annex II Clause 1.5.5, 1.5.6 and 1.5.7 of the Directive 94/9/EC is not covered by
this certificate.
CATALYTIC COMBUSTIBLE GAS SENSOR (CGS)
0539 II 2 G
Ex d IIC T3, T5 Gb
DEMKO 02 ATEX 131323X
T5 (Tamb = –40°C to +75°C)
T3 (Tamb = –55°C to +125°C)
FM
®
APPROVED
EN Standards:
EN 60079-0: 2009
EN 60079-1: 2007
EN 60079-29-1: 2007
10.1
C-1
95-8444
Special Conditions for Safe Use of CGS:
The CGS Combustible Gas Sensor is certified for use in following ambient temperatures:
- ambient temperature range –40°C to +75°C.
Coding: Ex d IIC T5 Gb
- ambient temperature range –55°C to +125°C.
Coding: Ex d IIC T3 Gb
The actual temperature range is marked on the sensor.
The performance ambient temperature rating is limited to –40°C to +75°C.
The CGS Combustible Gas Sensor can withstand repeated exposures to 125°C for periods up to 12 hours. It is
recommended that the sensor be replaced after maximum 500 hours of exposed to the 125°C temperature condition.
The CGS Combustible Gas Sensor must be used in conjunction with the before mentioned ATEX certified Detector
Electronics Corp. combustible gas detector control units for compliance with EN 60079-29-1 standards.
The CGS Combustible Gas Sensor must only be mounted into the enclosures of the Infiniti Gas Transmitter Model
U9500A Series, the Combustible Gas Transmitter Model 505 Series, the Digital Communication Unit EQ 22xxDCUEX
Series or the Sensor Termination Box Model STB Series.
The actual enclosure must provide a maximum measured reference pressure of 15 bar measured according to
EN 60079-1: 2007, §15.
The CGS Combustible Gas Sensor is to be installed in places where there is a low risk of mechanical damage.
SENSOR TERMINATION BOX (STB)
0539 II 2 G
Ex d IIC T4–T6 Gb
DEMKO 02 ATEX 131324X
T6 (Tamb = –55°C to +60°C).
T5 (Tamb = –55°C to +75°C).
T4 (Tamb = –55°C to +125°C).
IP66
FM
®
APPROVED
EN Standards:
EN 50270: 2006
EN 60079-0: 2009
EN 60079-1: 2007
EN 60529: 2001
EN 60529: 1991+A1: 2000
EN 60079-29-1: 2007
CE MARK
DIRECTIVE CONFORMITY
Low Voltage – 2006/95/EEC.
Electromagnetic compatibility – 2004/108/EC.
ATEX – 94/9/EC.
The U9500 Gas Transmitter was tested and found to be compliant with EN50270 when wired in conduit or with shielded
cable. All screen drains shall be terminated to the chassis.
The U9500 Gas Transmitter was tested to EN50270 and are compliant for Type 1 equipment. To meet Type 2
requirements, additional transient suppression is provided for installation.
10.1
C-2
95-8444
10.1
C-3
95-8444
Figure C1—Device Configuration for ATEX Approval
APPENDIX D
IECEx APPROVALS
INFINITI U9500 GAS TRANSMITTER
IECEx ULD 10.0009
Ex d IIC T5-T6 Gb
T6 (Tamb = –55°C to +60°C)
T5 (Tamb = –55°C to +75°C)
IP66
IEC Standards: IEC 60079-0: 2007
IEC 60079-1: 2007
IEC 60529, 2.1.ed.+Corr. 1:2003+2:2007
Read and understand instruction manual before operating.
To obviate the risk of hot spots and capacitor energy storage, the enclosure must not be opened, even when isolated,
when an explosive gas atmosphere is present.
All cable entry devices and blanking elements shall be certified in type of explosion protection flameproof enclosure
“d”, suitable for the conditions of use and correctly installed. Unused apertures shall be closed with suitable certified
blanking elements.
For ambient temperatures below –10°C and above +60°C use field wiring suitable for both minimum and maximum
ambient temperature.
CATALYTIC COMBUSTIBLE GAS SENSOR (CGS)
IECEx ULD 10.0001X
Ex d IIC T3, T5 Gb
T5 (Tamb = –40°C to +75°C)
T3 (Tamb = –55°C to +125°C)
IEC Standards:
IEC 60079-0: 2007
IEC 60079-1: 2007
CONDITION OF CERTIFICATION:
The CGS Combustible Gas Sensor is certified for use in following ambient temperatures:
- ambient temperature range –40°C to +75°C
Coding: Ex d IIC T5 Gb
- ambient temperature range –55°C to +125°C
Coding: Ex d IIC T3 Gb
The actual temperature range is marked on the sensor.
The CGS Combustible Gas Sensor can withstand repeated exposures to 125°C for periods up to 12 hours. It is
recommended that the sensor be replaced after maximum 500 hours of exposed to the 125°C temperature condition.
The CGS Combustible Gas Sensor must only be mounted into the enclosures of the Infiniti Gas Transmitter Model
U9500A Series, the Combustible Gas Transmitter Model 505 Series, the Digital Communication Unit EQ 22xxDCUEX
Series or the Sensor Termination Box Model STB Series.
10.1
D-1
95-8444
The actual enclosure must provide a maximum measured reference pressure of 15 bar measured according to
IEC 60079-1: 2007, §15.
The CGS Combustible Gas Sensor is to be installed in places where there is a low risk of mechanical damage.
SENSOR TERMINATION BOX (STB)
IECEx ULD 10.0007X
Ex d IIC T4–T6 Gb
T6 (Tamb = –55°C to +60°C)
T5 (Tamb = –55°C to +75°C)
T4 (Tamb = –55°C to +125°C)
IP66
IEC Standards:
IEC 60079-0: 2007
IEC 60079-1: 2007
IEC 60529, 2.1.ed.+Corr. 1:2003+2:2007
10.1
D-2
95-8444
APPENDIX D
ADDITIONAL APPROVALS
GOST-R
Infiniti Gas Transmitter Model U9500
VNIIFTRI GOST R Certificate of Compliance
1Ex d IIC
T6 (Tamb = –60°C to +50°C)
T5 (Tamb = –60°C to +65°C)
T4 (Tamb = –60°C to +75°C)
BRAZIL
Infiniti Gas Transmitter Model U9500
CEPEL 98.0013
Ex d IIC T5-T6 Gb IP66
T6 (Tamb = –55°C to +60°C)
T5 (Tamb = –55°C to +75°C)
IEC Standards: IEC 60079-0: 2007
IEC 60079-1: 2007
NOTE
All cable entry devices and blanking elements shall be Brazil certified in explosion protection, flameproof ‘d’,
suitable for the conditions and correctly installed with an IP66 rating. A screw or cover lock is provided for a
secondary means of fastening the cover.
10.1
E-1
95-8444
95-8444
Detector Electronics Corporation
6901 West 110th Street
Minneapolis, MN 55438 USA
X3301 Multispectrum
IR Flame Detector
PointWatch Eclipse®
IR Combustible Gas Detector
FlexVu® Universal Display
w/ GT3000 Toxic Gas Detector
Eagle Quantum Premier®
Safety System
T: 952.941.5665 or 800.765.3473
F: 952.829.8750
W: http://www.det-tronics.com
E: det-tronics@det-tronics.com
Det-Tronics, the DET-TRONICS logo, Infiniti, and Eclipse are registered trademarks or trademarks of Detector Electronics Corporation
in the United States, other countries, or both. Other company, product, or service names may be trademarks or service marks of others.
© Copyright Detector Electronics Corporation 2012. All rights reserved.