Yokogawa DO30G Sensor for Dissolved Oxygen Analyzer General Specifications
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
General
Specifications
DO30G
Sensor for Dissolved Oxygen Analyzer
GS 12J05D03-01E
n GENERAL
The DO30G sensor for dissolved oxygen is designed
for use in water treatment plants such as sewage
treatment works, effluent activated sludge process,
and potable water treatment. It is also effective in river
monitoring, intake protection, fish farming and other
fields where water quality is important.
This galvanic cell has a fast response time and good
long term stability, and features replaceable diaphragm
and cable for easy maintenance. A temperature sensor
is incorporated for compensation. The DO30G sensor
can be used in either the PB350G/PB360G floating ball
holders or in the DOX8HS submersion type holder.
DO30G Sensor
n FEATURES
DO Sensor with Long-term Stability and Short
Electrolyte Stabilization Time
• A DO sensor that uses a special electrolyte, shortening the electrolyte stabilization time and realizing
stable measurements over a long period of time.
• Measuring range of 0 to 20 mg/L
• A membrane that is easily replaceable by anyone
because of the cartridge type
• A membrane 50 mm thick that is hard to tear and has
reduced influence from air bubbles
PB350G Floating ball holder
n SYSTEM
Refer to GS 12J5D2-E for dissolved oxygen converter DO402G, GS 12J05C01-01E for 2-wire transmitter DO202,
GS 12A01A03-01EN for 2-wire liquid analyzer FLXA202, GS 12A01A02-01E for 2-wire liquid analyzer FLXA21 and
GS 12J05C02-00E for holders
DO Sensor
• DO30G
Select pin terminal
of DO30G sensor for
DO402G, DO202.
Accessories
• DOX8A
Holder, Holder with Cleaner
• Guide pipe
PH8HG
DO Converter
• Floating ball holder
PB350G
PB360G
• DO402G, DO202
or FLXA202/FLXA21
Jet cleaner
• WTB10-DO3, -DO4
• Submersion
type holder
DOX8HS
Without
cleaner
Terminal Box
• Suspension holder
HH350G
Note: “DO” of DO sensor means dissolved oxygen.
CAUTION
Installation Location of Holders (Guide Pipe, Submersion Type, etc)
The holder should be used in a place that is as vibration free as possible.
Using the holder in a place where it is affected by vibration, may result in damage to the holder.
Yokogawa Electric Corporation
2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 Japan
Tel.: 81-422-52-5617 Fax.: 81-422-52-6792
GS 12J05D03-01E
©Copyright May. 2007 (YK)
5th Edition: Sep. 2015 (YK)
2
n GENERAL SPECIFICATIONS
GENERAL SPECIFICATIONS of DO30G SENSOR
Dissolved Oxygen:
0 to 20 mg/L (ppm)
Liquid Temperature:
0 to 40˚C
Liquid Pressure:
0 to 100 kPa
Flow speed: 20 cm/s or more
Sensor for Temperature Compensation: Pt1000
Wetted Part Material:
rigid polyvinyl chloride, stainless steel,
fluorinated ethylenepropylene, nitrile rubber, heat-resistant soft polyvinyl chloride,
and polycarbonate
Cable Length: 3 m, 5 m, 10 m, 15 m and 20 m
Weight: Approx. 0.3 kg + 0.12 × N kg
N: cable length
(ex. for cable length of 5 m, approx. 0.9 kg)
CHARACTERISTICS of DO30G SENSOR
Repeatability: 0.1 mg/L or 3% FS, whichever is greater
(including sensor)
Temperature Compensation Error:
Within ±3% FS (including sensor) for a
±5˚C change in the range of 0 to 40˚C
Response Time: Within 2 minutes
(90% response)(including sensor)
WTB10 Terminal Box
Used when DO converter is installed at a distance
from the DO sensor.
Ambient temperature: -10 to 50˚C
Construction: JIS waterproof
Case material: Fiberglass reinforced polycarbonate resin
Electrical connections
DO sensor side: JIS A8 watertight plastic gland
DO converter side: JIS A15 watertight plastic gland
with cable (max. 40 m)
Conduit adaptor (optional) available
Case color:
Grayish green
(Munsell 2.5GY5.0/1.0)
Weight: Body 0.5 kg Mounting hardware: 0.7 kg
DOX8A Parts Set for Maintenance
A set of zero-adjusting reagent and maintenance parts
Zero adjusting reagent
(sodium sulphite 500 g)
1 bottle
Membrane assembly
(for membrane thickness of 50 mm)
3 sets
Electrolyte for sensor (50 mL)
1 bottle
Syringe for replacing electrolyte
1
Polish for silver electrode (30 g)
1 bottle
Polyethylene beaker (200 mL)
1
DOX8W Calibration Set (optional)
This is necessary if the span calibration is to be done
using a saturated dissolved oxygen solution. It is not
necessary for air calibration.
Air pump, stirrer, stirring element, bubbler, clamp,
beaker, and thermometer.
Note: The calibration set can be used in common
regardless of the type of holder.
n MODEL AND SUFFIX CODES
1. DO Sensor
Model
Suffix Code Option Code
DO30G
•••••••••••••••
••••••••••••••• Dissolved Oxygen Sensor
―
-NN
••••••••••••••• Always -NN
Membrane -50
thickness
••••••••••••••• 50 µm
-03
-05
-10
-15
-20
Cable length
••••••••••••••• 3 m
••••••••••••••• 5 m
••••••••••••••• 10 m
••••••••••••••• 15 m
••••••••••••••• 20 m
-PN ••••••••••••••• Pin terminal (*1)
-FK ••••••••••••••• Fork terminal
-FL ••••••••••••••• M4 ring terminal (*2)
Select pin terminal for DO402G, DO202, and
FLXA202/FLXA21.
When terminal box is used, select WTB10-DO3.
Used to connection to FLXA202/FLXA21.
When terminal box is used, select WTB10-DO4.
Cable terminal
*1:
*2:
2. Terminal Box
Model
WTB10
Suffix Code
Option
Code
•••••••••••••••••
••••••••• Terminal box
-DO3
Combinned
system
-DO4
-NN
―
••••••••• Always -NN
-00
-05
-10
-20
-30
-40
Specify cable
length
Description
••••••••• For DO402G,DO202, and
FLXA202/FLXA21 (*1)
••••••••• For FLXA202/FLXA21 (*2)
•••••••••
•••••••••
•••••••••
•••••••••
•••••••••
•••••••••
No cable
5m
10 m
20 m
30 m
30 m
Mounting hardware /P
••••••••• Pipe mounting hardware
/W
Option
••••••••• Wall mounting hardware
Conduit work /AWTB ••••••••• G 1/2
adaptor /ANSI ••••••••• 1/2NPT
*1: Used for pin terminals, and cable with pin terminals.
*2: Used for M4 screw terminals, and cable with M4
ring terminals.
3. Accessories (parts set for maintenance)
Model
Suffix Code Option Code
DOX8A
Sensor
-M
••••••••••••••• For membrane replacing
type sensor
*B
―
Description
••••••••••••••• Parts set for maintenance
(for membrane
thickness of 50 µm)
••••••••••••••• Style B
4. Calibration Set
Model
Suffix Code Option Code
DOX8W *A
••••••••••••••
Description
Calibration set
5. Spare Parts
Part Name
Part No.
Membrane and
K9171HK
electrolyte for sensor
All Rights Reserved. Copyright © 2007, Yokogawa Electric Corporation
Description
Remarks
3 membrane assemblies
(thickness of 50 µm)
50 mL of electrolyte
Syringe for replacing electrolyte
Zero adjusting reagent L9920BR
Sodium sulfite
Polisher
For polishing silver electrode 30g
K9088PE
GS 12J05D03-01E
500g
5th Edition: Nov. 13,2015-00
3
n DIMENSIONS
DO Sensor DO30G
Unit: mm
Dissolved-oxygen sensor
with fork terminals
Dissolved-oxygen sensor
with pin terminals
11 12 14 13 15 16
11 12 14 13 15 16
Pin terminals
Model and Suffix Codes
DO30G - NN - 50 - 03 - PN
DO30G - NN - 50 - 05 - PN
DO30G - NN - 50 - 10 - PN
DO30G - NN - 50 - 15 - PN
DO30G - NN - 50 - 20 - PN
約80
Dissolved-oxygen sensor
with ring terminals
Weight (kg)
Approx. 0.6
Approx. 0.8
Approx. 1.4
Approx. 2.0
Approx. 2.6
L
3 000
5 000
10 000
15 000
20 000
Weight (kg)
Approx. 0.6
Approx. 0.8
Approx. 1.4
Approx. 2.0
Approx. 2.6
L
3 000
5 000
10 000
15 000
20 000
Weight (kg)
Approx. 0.6
Approx. 0.8
Approx. 1.4
Approx. 2.0
Approx. 2.6
Fork terminals
Model and Suffix Codes
DO30G - NN - 50 - 03 - FK
DO30G - NN - 50 - 05 - FK
DO30G - NN - 50 - 10 - FK
DO30G - NN - 50 - 15 - FK
DO30G - NN - 50 - 20 - FK
11 12 14 13 15 16
ø6.5
L
3 000
5 000
10 000
15 000
20 000
L
Ring terminals
ø26
Model and Suffix Codes
DO30G - NN - 50 - 03 - FL
DO30G - NN - 50 - 05 - FL
DO30G - NN - 50 - 10 - FL
DO30G - NN - 50 - 15 - FL
DO30G - NN - 50 - 20 - FL
38.5 46
10
Terminal asign
13
15
11
12
14
16
108
F02.ai
IE
RE
T1
T2
Shield
Liquid earth
ø30
ø36
Terminal Box WTB10
65
4-M5 screws, depth 9
(for mounting)
46
Converter cable (Extension cable) inlet (to be drilled for wiring)
with cable gland (applicable cable o.d.: ø9 to 12 mm)
Sensor cable inlet (to be drilled for wiring)
with cable gland (applicable cable o.d.: ø5.5 to 7 mm)
22
75 ±0.2
178
59
126
Unit: mm
70±0.2
All Rights Reserved. Copyright © 2007, Yokogawa Electric Corporation
F03.ai
GS 12J05D03-01E
5th Edition: Sep. 18,2015-00
4
Detailed drawing of cable gland for terminal box
Packing
Retaining Claw
49
Approx. 26
Approx. 55
Unit: mm
G1/2 (/AWTG)
1/2NPT (/ANSI)
Adapter for Conduit
F04.ai
Please use the watertight cable gland of the attachment.
If metallic conduit is directly connected with the case,
it becomes the factor of the measurement error.
(b) Watertight plastic cable gland
with conduit adaptor
(a) Watertight plastic cable gland (St'd)
Dedicated Extension Cable with WTB10 (Not supplied if suffix code “-00” is selected)
WTB10-DO3
Approx. ø6.5
Approx. ø9.5
WTB10-DO4 WTB10-DO3
Unit: mm
WTB10-DO4
14 (S)
1 4
15 (RE)
1 5
13 (IE)
1 3
16 (SE)
1 6
1 6
11 (T1)
1 1
1 1
11 (T1)
12 (T2)
1 2
1 2
12 (T2)
40
80
80
40
45
1 4
14 (S)
1 5
15 (RE)
1 3
13 (IE)
16 (SE)
45
80
80
L
Model and Suffix Code
WTB10 - DO3 - NN - 05, WTB10 - DO4 - NN - 05
WTB10 - DO3 - NN - 10, WTB10 - DO4 - NN - 10
WTB10 - DO3 - NN - 20, WTB10 - DO4 - NN - 20
WTB10 - DO3 - NN - 30, WTB10 - DO4 - NN - 30
WTB10 - DO3 - NN - 40, WTB10 - DO4 - NN - 40
L (mm)
Approx. 5000
Approx. 10000
Approx. 20000
Approx. 30000
Approx. 40000
Weight : Approx. 0.12 kg/m
F05.ai
Pipe Mounting Bracket (Optional) (Option Code: /P)
100
70±0.2
Unit: mm
4 - ø5.5 holes
(254)
75±0.2
8
4-M5 screws, depth 9
47
70
203
U-bolt, washer, nut
70
4 - ø10 holes
All Rights Reserved. Copyright © 2007, Yokogawa Electric Corporation
(65)
48
Nominal 50A
(outer diameter ø60.5)
pipe
Weight: approximately 700 g
GS 12J05D03-01E
F06.ai
5th Edition: Sep. 18,2015-00
5
Wall Mounting Bracket (Optional) (Option Code: /W)
Unit: mm
4-ø5.5 holes
100
70±0.2
(236)
15
27
185
75±0.2
8
4-M5 screws, depth 9
70
(65)
3-ø10 holes
13
Weight: approximately 300 g
F07.ai
n WIRING DIAGRAM EXAMPLE
DO402G
*1
Output signal
(4 to 20 mA DC or 0 to 20 mA DC)
61
62
65
Output signal
(4 to 20 mA DC or 0 to 20 mA DC)
66
*1
Contact input
(cleaning start command)
mA 1
mA 2
22
N
2
G
3
S1
13
11
11 T1
12
12 T2
31
32
33
41
S2
42
43
23 SCREEN
17
Jumper
13 IE
WTB10 Terminal box *4
1
63 SCREEN
21
*7
L
51
S3
14
14 Shield
15
15 RE
16
16 Liquid earth *6
52
53
71
S4
72
73
Note : External wiring connection terminal size is for ø2 pin.
DO30G
DO sensor
F08.ai
*3
Power supply
*3 C
NC
NO
Contact
output S1
C
NC
NO
*3 C
NC
NO
C
NC
NO
Contact
output S2
High and low
alarms or HOLD
Contact output S3 (cleaning,
HOLD or high and low alarms)
Contact output S4/FAIL
(failure or high and low alarms)
Grounding terminal (M4 screw)
*2
Protective grounding (100Ω or less)
*1: Always use a shielded cable with an OD of 6 to 12 mm.
*2: Be sure to ground the DO converter case grounding terminal (grounding resistance of 100Ω or less).
*3: Always use a cable with an OD of 6 to 12 mm.
*4: Terminal box is used only where DO converter is installed long distance from DO sensor. (ordinary not needed)
*5: This cable is specified in the Basic Code for the WTB10.
*6: Liquid earth to detect membrane failure. This function is available when using PB350G/PB360G float type holder.
Connect only when using this function.
*7: Refer to GS of “DO402G(GS 12J05D02-00E)" when contact input of DO402G is used.
All Rights Reserved. Copyright © 2007, Yokogawa Electric Corporation
GS 12J05D03-01E
5th Edition: Sep. 18,2015-00
6
Inquiry Specifications Sheet for Dissolved Oxygen Analyzer
For inquiries on the Yokogawa dissolved oxygen analyzer, please tick (ü) the appropriate box (¨) and write down the
relevant information in the blanks.
1. General
Name of your company:
Name of person
:
Dept. or sect :
Name of plant
:
Measuring point
:
Purpose of use
: ¨ Indication ¨ Recording ¨ Alarm ¨ Control
Power supply
:
Hz
V¨AC
¨ DC
2. Measurement Conditions
(1) Liquid temperature :
to
, Normal
(2) Liquid pressure
:
to
, Normal
(3) Flow speed
:
to
, Normal
(4) Name of liquid to be measured :
(5) Components of liquid to be measured :
(6) Other conditions
:
(Phone No.
[°C]
[kPa]
[m/s]
3. Installation environment
(1) Ambient temperature : approx.
[°C]
(2) Location
: ¨ Outdoors ¨ Indoors
(3) Other information
:
4. Requirements
(1) Measurement range : ¨ 0 to 20 mg/L ¨
(2) System component selection: ¨ Sensor ¨ Holder ¨ Converter ¨ Cleaning system ¨ Calibration set
¨ Parts set for maintenance ¨ Terminal Box
(3) Length of sensor cable : ¨ 3 m ¨ 5 m ¨ 10 m ¨ 15 m ¨ 20 m
(4) Type of holder
: ¨ Guide pipe ¨ Submersion ¨ Floating ball ¨ Suspension
(5) Cleaning method
: ¨ No cleaning ¨ Jet cleaning
(6) Other requirements :
All Rights Reserved. Copyright © 2007, Yokogawa Electric Corporation
Subject to change without notice.
GS 12J05D03-01E
5th Edition: Sep. 18,2015-00
)