Teledyne Analytical Instruments OPERATING INSTRUCTIONS FOR Model 3020M Percent Paramagnetic Oxygen Analyzer
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Percent Paramagnetic Oxygen Analyzer
OPERATING INSTRUCTIONS FOR
Model 3020M
Percent Paramagnetic
Oxygen Analyzer
DANGER
HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING
SYSTEM.
PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.
HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST
FOR A TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.
ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING. BEFORE
CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/
MANAGER.
Teledyne Analytical Instruments
P/N M69993
12/18/98
ECO:#98-0569
i
Model 3020M
Copyright © 1998 Teledyne Analytical Instruments
All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer
language in whole or in part, in any form or by any means, whether it be electronic,
mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of
Teledyne Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91749-1580.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us free
from defects of material and of construction, and that our liability shall be limited to
replacing or repairing at our factory (without charge, except for transportation), or at
customer plant at our option, any material or construction in which defects become
apparent within one year from the date of shipment, except in cases where quotations or
acknowledgements provide for a shorter period. Components manufactured by others bear
the warranty of their manufacturer. This warranty does not cover defects caused by wear,
accident, misuse, neglect or repairs other than those performed by Teledyne or an authorized service center. We assume no liability for direct or indirect damages of any kind and
the purchaser by the acceptance of the equipment will assume all liability for any damage
which may result from its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our
apparatus, and to make any alterations in the dimensions, shape or weight of any parts, in
so far as such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its user, and serves as a tool by
which valuable data can be gathered. The information provided by the instrument may
assist the user in eliminating potential hazards caused by his process; however, it is
essential that all personnel involved in the use of the instrument or its interface, with the
process being measured, be properly trained in the process itself, as well as all instrumentation related to it.
The safety of personnel is ultimately the responsibility of those who control process
conditions. While this instrument may be able to provide early warning of imminent danger,
it has no control over process conditions, and it can be misused. In particular, any alarm or
control systems installed must be tested and understood, both as to how they operate and
as to how they can be defeated. Any safeguards required such as locks, labels, or redundancy, must be provided by the user or specifically requested of Teledyne at the time the
order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions. The
purchaser is responsible for the training of personnel, for providing hazard warning
methods and instrumentation per the appropriate standards, and for ensuring that hazard
warning devices and instrumentation are maintained and operated properly.
Teledyne Analytical Instruments, the manufacturer of this instrument, cannot
accept responsibility for conditions beyond its knowledge and control. No statement
expressed or implied by this document or any information disseminated by the manufacturer or its agents, is to be construed as a warranty of adequate safety control under the
user’s process conditions.
ii
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Specific Model Information
The instrument for which this manual was supplied may incorporate one or
more options not supplied in the standard instrument. Commonly available
options are listed below, with check boxes. Any that are incorporated in the
instrument for which this manual is supplied are indicated by a check mark in the
box.
Instrument Serial Number: _______________________
Options Included in the Instrument with the Above Serial Number:
❑ 3020M-C:
In addition to all standard features, this model also has
separate ports for zero and span gases, and built-in control
valves. The internal valves are entirely under the control of
the 3020M electronics, to automatically switch between
gases in synchronization with the analyzer’s operations
❑ 3020M-F:
Includes flame arrestors for Group C and D service.
❑ 3020M-G:
Includes flame arrestors for Groups C and D service, plus
gas control valves as in –C option, above
❑ 3020M-H:
Includes flame arrestors for Group B (hydrogen) service.
❑ 3020M-I:
Includes flame arrestors for Group B (hydrogen) service,
plus gas control valves as in –C option, above.
❑ 3020M-M:
4-20 mA dc Signal and Range ID outputs (in addition to
the standard dc voltage outputs.
Teledyne Analytical Instruments
i
Model 3020M
Table of Contents
1 Introduction
1.1 Overview ........................................................................ 1-1
1.2 Typical Applications ....................................................... 1-1
1.3 Main Features of the Analyzer ....................................... 1-1
1.4 Model Designations ....................................................... 1-2
1.5 Operator Interface .......................................................... 1-2
1.5.1 UP/DOWN Switch .................................................. 1-4
1.5.2 ESCAPE/ENTER Switch ....................................... 1-4
1.5.3 Displays ................................................................. 1-5
1.6 Recognazing Difference Between LCD & VFD.............. 1-5
1.7 Rear Panel (Equipment Interface) .................................. 1-5
1.7.1 Electrical Connector Panel .................................... 1-5
1.7.2 Gas Connector Panel ............................................. 1-7
2 Operational Theory
2.1 Introduction .................................................................... 2-1
2.2 Paramagnetic Sensor..................................................... 2-1
2.2.1 Principles of Operation ............................................ 2-1
2.3 Cross Interference .......................................................... 2-3
2.4 Sample System .............................................................. 2-5
2.5 Electronics and Signal Processing ................................ 2-6
2.6 Temperature Control ...................................................... 2-6
3 Installation
3.1 Unpacking the Analyzer ................................................. 3-1
3.2 Mounting the Analyzer ................................................... 3-1
3.3 Electrical Connections ................................................... 3-3
3.3.1 Primary Input Power ............................................... 3-4
3.3.2 Fuse Installation..................................................... 3-4
3.3.3 Analog Outputs ...................................................... 3-4
3.3.4 Alarm Relays ......................................................... 3-7
3.3.5 Digital Remote Cal Inputs ...................................... 3-8
3.3.6 Range ID Relays ................................................... 3-9
3.3.7 Network I/O ............................................................ 3-9
3.3.8 RS-232 Port ........................................................... 3-9
3.3.9 Remote Sensor and Solenoid Valves .................... 3-11
3.4 Gas Connections .......................................................... 3-12
3.5 Testing the System ........................................................ 3-13
iv
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
4 Operation
4.1 Introduction .................................................................... 4-1
4.2 Using the Controls ......................................................... 4-1
4.2.1 Mode/Function Selection ....................................... 4-2
4.2.1.1 Analysis Mode ............................................... 4-2
4.2.1.2 Setup Mode ................................................... 4-2
4.2.2 Data Entry .............................................................. 4-4
4.2.2.1 ENTER .......................................................... 4-4
4.2.2.2 ESCAPE ........................................................ 4-4
4.3 The AUTO-CAL Function ............................................... 4-5
4.4 The PWD Function ........................................................ 4-5
4.4.1 Entering the Password ........................................... 4-6
4.4.2 Installing or Changing the Password ..................... 4-7
4.5 The LOGOUT Function .................................................. 4-8
4.6 The VERSION Screen ................................................... 4-8
4.7 The SELF TEST Function.............................................. 4-9
4.8 The SPAN Functions ..................................................... 4-9
4.8.1 Span Cal ................................................................ 4-10
4.8.1.1 Auto Mode Spanning ..................................... 4-10
4.8.1.2 Manual Mode Spanning ................................. 4-11
4.9 The CAL Function .......................................................... 4-11
4.10 The ALARMS Function .................................................. 4-13
4.11 The RANGE Function .................................................... 4-15
4.11.1 Setting the Analog Output Ranges......................... 4-16
4.11.2 Automatic Ranging ................................................ 4-16
4.11.3 Fixed Range Analysis ............................................ 4-16
4.12 The CONTRAST Function ............................................. 4-17
4.13 The STANDBY Function ................................................ 4-17
4.14 The Analysis Mode ........................................................ 4-18
4.15 The Sensor Function ..................................................... 4-18
Maintenance
5.1
5.2
5.3
5.4
5.5
5.6
Routine Maintenance ..................................................... 5-1
Major Internal Components ............................................ 5-1
Sensor Replacement ..................................................... 5-2
Fuse Replacement......................................................... 5-2
System Self Diagnostic Test ........................................... 5-3
Output Goes Negative .................................................... 5-4
Teledyne Analytical Instruments
v
Model 3020M
Appendix
A-1 Specifications ................................................................ A-1
A-2 Recommended 2-Year Spare Parts List ......................... A-3
A-3 Drawing List ................................................................... A-4
vi
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Introduction 1
Introduction
1.1
Overview
The Teledyne Analytical Instruments Model 3020M Percent Paramagnetic Oxygen Analyzer is a versatile microprocessor-based instrument for
detecting oxygen in a variety of gases. This manual covers the Model
3020M, percent oxygen, explosion-proof, bulkhead-mount units only.
1.2
Typical Applications
A few typical applications of the Model 3020M are:
• Monitoring inert gas blanketing
• Air separation and liquefaction
• Chemical reaction monitoring
• Semiconductor manufacturing
• Petrochemical process control
• Quality assurance
• Gas analysis certification.
1.3
Main Features of the Analyzer
The Model 3020M Percent Oxygen Analyzer is sophisticated yet
simple to use. The main features of the analyzer include:
•
A 2-line alphanumeric display screen, driven by microprocessor
electronics, that continuously prompts and informs the operator.
•
High resolution, accurate readings of oxygen content from low %
levels through 100%. Large, bright, meter readout.
•
Stainless steel sample system.
Teledyne Analytical Instruments
1-1
1 Introduction
Model 3020M
•
Versatile analysis over a wide range of applications.
•
Microprocessor based electronics: 8-bit CMOS microprocessor
with 32 kB RAM and 128 kB ROM.
•
Three user definable output ranges (from 0-5 % through 0-100
%) allow best match to users process and equipment.
•
Auto Ranging allows analyzer to automatically select the proper
preset range for a given measurement. Manual override allows
the user to lock onto a specific range of interest.
•
Two adjustable concentration alarms and a system failure alarm.
•
Extensive self-diagnostic testing, at startup and on demand, with
continuous power-supply monitoring.
•
RS-232 serial digital port for use with a computer or other digital
communication device.
•
Analog outputs for concentration and range identification.
(0-1 VDC standard, and isolated 4–20 mADC)
1.4
Model Designations
3020M:
Standard model.
3020M-C:
In addition to all standard features, this model also has
separate ports for zero and span gases, and built-in control
valves. The internal valves are entirely under the control of
the 3020M electronics, to automatically switch between gases
in synchronization with the analyzer’s operations
3020M-F:
3020M-G:
3020M-H:
3020M-I:
Includes flame arrestors for Groups C and D.
Includes flame arrestors for Groups C and D, & -C option.
Includes flame arrestors for Group B (Hydrogen service).
Includes flame arrestors for Group B, & -C option.
n.
1.5
Operator Interface
All controls and displays on the standard 3020M are accessible from
outside the housing. The instrument has two simple operator controls. The
operator has constant feedback from the instrument through an alphanumeric
display, a digital oxygen meter, and a sample flow meter. The displays and
1-2
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Introduction 1
controls are described briefly here and in greater detail in chapter 4. See
Figure 1-1.
Through Side Port
DIGITAL METER
VFD SCREEN
Through Side Port
Through Side Port
DOWN/UP
CONTROL
Through Side Port
ESCAPE/ENTER
CONTROL
SCCM
Figure 1-1: Model 3020M Controls, Indicators, and Connectors
Teledyne Analytical Instruments
1-3
1 Introduction
Model 3020M
1.5.1 UP/DOWN Switch
Functions: The UP/DOWN switch is used to select the function to be
performed. Choose UP or DOWN to scroll through the following list of
twelve functions:
• Auto-Cal Set up an automatic calibration sequence.
• PWD
Install a password to protect your analyzer setup.
• Logout Locks Setup Mode.
• Version Displays model and version of analyzer.
• Self-Test Runs internal diagnostic program, displays results.
• Span
Span calibrate the analyzer.
• Zero
Zero calibrate the analyzer.
• Alarms Set the alarm setpoints and attributes.
• Range
Set up the 3 user definable ranges for the instrument.
Contrast Function is DISABLED
• Contrast Adjust LCD contrast.
(Refer to Section 1.6)
• Standby Leave analyzer powered, but no outputs or displays.
WARNING: THE POWER CABLE MUST BE DISCONNECTED TO
FULLY REMOVE POWER FROM THE INSTRUMENT.
CAUTION: VOLTAGE MAY BE PRESENT ON ALARM CONTACTS WHEN LINE POWER IS DISCONNECTED.
Subfunctions: Once a Function is entered, the UP/DOWN switch is
used to select between any subfunctions displayed on the VFD screen.
Parameter values: When modifiable values are displayed on the
VFD, the UP/DOWN switch can be used to increment or decrement the
values.
1.5.2 ESCAPE/ENTER Switch
Data Entry: The ESCAPE/ENTER switch is used to input data, from
the alphanumeric VFD screen into the instrument:
•
Escape Moves VFD display back to the previous screen in a
series. If none remains, returns to the Analyze screen.
With subfunction selected, moves VFD back through
items on screen, to first item, then moves VFD to
previous display.
•
1-4
Enter
With a Subfunction Selected: Moves VFD on to the
next screen in a series. If none remains, returns to the
Analyze screen.
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Introduction 1
With a Value Selected: Enters the value into the
analyzer as data. Advances VFD to next operation.
(See Chapter 4 for details.)
1.5.3 Displays
Digital Meter Display: The meter display is a LED device that
produces large, bright, 7-segment numbers that are legible in any lighting. It
produces a continuous readout from 0-100 %. It is accurate across all
analysis ranges without the discontinuity inherent in analog range switching.
Alphanumeric Interface Screen: The backlit VFD screen is an easyto-use interface from operator to analyzer. It displays values, options, and
messages that give the operator immediate feedback.
Flowmeter: Monitors the flow of gas past the sensor. Readout is 100 to
1000 standard cc per minute (cc/min x 100) valid for air or nitrogen.
1.6
Recognizing Difference Between LCD &
VFD
LCD has GREEN background with BLACK characters. VFD has
DARK background with GREEN characters. In the case of VFD - NO
CONTRAST ADJUSTMENT IS NEEDED.
1.7
Equipment Interface
1.7.1 Electrical Connector Panel
The electrical connector panel, shown in Figure 1-2, contains the
electrical connections for external inlets and outlets. The connectors are
described briefly here and in detail in the Installation chapter of this manual.
CAUTION: The power cable must be disconnected to fully
remove power from the instrument.
CAUTION: VOLTAGE MAY BE PRESENT ON ALARM CONTACTS WHEN LINE POWER IS DISCONNECTED.
Teledyne Analytical Instruments
1-5
1 Introduction
Model 3020M
100-240 V
50/60Hz
3.0A Max.
Figure 1-2: Electrical Connector Panel
1-6
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Introduction 1
Electrical Connections: The electrical connections on the electrical
connector panel are described briefly here, and in more detail in chapter 3
Installation.
•
Power Connection
100-240 VAC, 50 or 60 Hz.
•
Analog Outputs
0-1 VDC concentration plus 0-1 V dc
range ID. Additional, isolated 4-20 mA
DC plus 4-20 mADC range ID
available.
•
Alarm Connections
2 concentration alarms and 1 system
alarm.
•
RS-232 Port
Serial digital concentration signal output
and control input.
•
Remote Valves
Used for controlling external solenoid
valves, if desired.
•
Remote Span/Zero
Digital inputs allow external control of
analyzer calibration.
•
Calibration Contact
To notify external equipment that
instrument is being calibrated and
readings are not monitoring sample.
•
Range ID Contacts
Four separate, dedicated, range relay
contacts. Low, Medium, High, Cal.
•
Network I/O
Serial digital communications for local
network access. For future expansion.
Not implemented at this printing.
1.7.2 Gas Connector Panel
The gas connector panel, shown in Figure 1-3, contains the gas connections for external inlets and outlets. Those that are optional are shown
shaded in the figure. The connectors are described briefly here and in detail
in the Installation chapter of this manual.
Teledyne Analytical Instruments
1-7
1 Introduction
Model 3020M
SCCM
Figure 1-3: Model 3020M Gas Connector Panel
•
Gas Sample Inlet and Outlet
One inlet and one exhaust out.
Optional:
•
Calibration Gas Ports Separate fittings for zero, span and
sample gas input, plus internal valves for
automatically switching the gases in
sync with the 3020M electronics.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the
Model 3020M is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.
1-8
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Operational Theory 2
Operational Theory
2.1
Introduction
The analyzer is composed of three subsystems:
1. Paramagnetic Sensor
2. Sample System
3. Electronic Signal Processing, Display and Control
The sample system is designed to accept the sample gas and transport it
through the analyzer without contaminating or altering the sample prior to
analysis. The Paramagnetic Sensor is an electromechanical device that
translates the amount of oxygen present in the sample into an electrical
signal. The electronic signal processing, display and control subsystem
simplifies operation of the analyzer and accurately processes the sampled
data. The microprocessor controls all signal processing, input/output and
display functions for the analyzer.
2.2
Precise Paramagnetic Sensor
2.2.1 Principles of Operation
The heart of the 3020M is a paramagnetic type oxygen sensor that is
maintanance free and has a long lifetime. Oxygen has a very high magnetic
sucseptibility compared to other gases and thus displays a particularly paramagnetic behaviour. A small glass dumbbell filled with nitrogen and rotating on a taut platinum wire is suspended in an inhomogneous magnetic field.
This glass dumbbell is diamagnetic and tends to rotate out of the magnetic
field. The strength of the resulting torque determined by an susceptibility of
the sample gas. This torque is compensated for by a counter torque induced
by an electrically charged platinum coil on the dumbbell. The zero position
of the dumbbell is controlled by means of an optical system consisting of a
light source, a mirror at the dumbbell axis and a pair of detectors. The
difference between the compensating currents required to bring the dumbTeledyne Analytical Instruments
2-1
2 Operational Theory
Model 3020M
bell to the zero position in the presence of zero gas (i.e. no O2 present) or of
sample gas is proportional to the partial pressure of oxygen in the sample
gas.
The sensor enclosure is temperature controlled to 55 degrees centigrade
to insure that the magnetic susceptibility of oxygen in the sample is not
affected by the ambient temperature. The measuring cell has a measuring
volume of approximately 3 cm3
Sensor
(side view shown)
The Paramagnetic sensor enclosure holds not only the sensing elements,
but the temperature controller electronics, heating elements, sensor electronics, and preamplifier. The Electronics and heating elements require a separate
power source, from the rest of the 3020M capable of delivering 1.5 amps
approximately at 24 volts dc. The output of the sensor is roughly calibrated
to be 0 to 1 volt DC for the the range of 0 to 100 % 02. The true calibration
of the sensor is carried out by the microprocessor as described later in chapter 4. The electrical interconnections to the sensor are done through a 15 pin
D connector. Some signals from the sensor are not connected. They are only
useful for troubleshooting, by trained personnel, as test points.
2-2
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Operational Theory 2
Sensor
(rear view shown)
Pin out:
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
2.3
-15Vdc test pin (Not connected)
+15Vdc test pin (Not connected)
Measuring ground
Not used
Preamplifier output ~0- 1 Vdc
Not used
24 Vdc return, power ground
Collector of transistor switching heating element (Not connected)
Fault signal: <0 Vdc= OK, >4.6 Vdc = Fault (Not connected)
Not used
Negative reference voltage (Not connected)
Positive reference voltage (Not connected)
Nominal temperature voltage signal (Not connected)
Actual temperature voltage signal (Not connected)
+24 Vdc power.
Cross Interference
As mentioned previously, the selectivity of the measuring system is
based on the extraordinarily high magnetic susceptibility of Oxygen compared to other gases. In most cases the presence of other gases can be ignored but significant errors can occur when the sensor is calibrated with a
mixture of oxygen and nitrogen and the sample gas consists mainly of other
gases with considerable magnetic susceptibility. In this case, the reading
shows a measured value even if the sample gas contains no Oxygen. It is
actually displaying the cross sensitivity to another gas.
The following table shows the cross sensitivity of some gases when
changing from pure nitrogen to 100% of one of the gases listed.
Teledyne Analytical Instruments
2-3
2 Operational Theory
Model 3020M
Gas
Cross
Sensitivity
in vol. %
Gas
Cross
Sensitivity
in vol. %
Acetylene C2H2
Allene C3H4
Ammonia NH3
Argon Ar
Bromine Br2
1.2 Butadiene C4H6
1.3 Butadiene C4H6
n-Butane C4H10
I-Butane C4H8
cis 2-Butane C4H8
trans 2-Butane C4H8
Carbon dioxide CO2
Carbon monoxide CO
Chlorine Cl2
Cyclo hexane C6H12
Ethane C2H6
Ethylene C2H4
Helium He
n-Heptane C7H16
n-Hexane C6H14
Hydrogen H2
Hydrogen bromide HBr
-0.24
-0.44
-0.26
-0.22
-1.30
-0.49
-0.49
-1.11
-085
-0.89
-0.92
-0.27
+0.06
-0.77
-1.56
-0.43
-0.26
+0.30
-2.10
-1.70
+0.24
-.0.61
Hydrogen chloride HCI
Hydrogen fluoride HF
Hydrogen iodide HI
Hydrogen sulphide H2S
Iodine I
Isobutane C4H10
Isopantane C5H12
Krypton Kr
Laughing gas N2O
Methane CH4
Neon Ne
Neopentane C5H12
Nitric acid HNO3
Nitrogen dioxide NO2
Nitrous oxide NO
n-Octane C8H18
n-Pentane C5H12
Propane C3H8
Propylene C3H6
Vinyl chloride
Water H2O
Xenon Xe
-0.30
+0.10
-1.10
-0.39
-2.40
-1.11
-1.49
-0.51
-0.20
-0.20
+0.13
-1.49
+0.43
+28.00
+40.00
-2.50
-1.45
-0.86
-0.55
-0.63
-0.02
-0.95
With gas mixtures the components are weighted according to their
proporational volumes.
The cross interference with a gas mixture can be determined in advance if the
concentration of the individual background gases is known.
Example:
Cross sensitivity calculation for a gas mixture:
The gas to be measured consists of 10 % CO2, 40 % of N2, and 50%
of Ar.
C02: -0.27 x 10% = -0.027
Ar: -0.22x5O%= -0.11
TOTAL CROSS INTERFERENCE -0.137
Calculation of the zero value:
2-4
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Operational Theory 2
(0% 02) - (Cross Interference Value) = Zero Point Value
0% -(-0.137) = +0.137%02
2.4
Sample System
The sample system delivers gases to the sensor from the analyzer gas
panel inlets. Depending on the mode of operation either sample or calibration
gas is delivered.
The Model 3020M sample system is designed and fabricated to ensure that
the oxygen concentration of the gas is not altered as it travels through the
sample system. The sample encounters almost no dead space. This minimizes
residual gas pockets that can interfere with low percent range analysis.
The sample system for the standard instrument incorporates ¼ inch tube
fittings for sample inlet and outlet connections at the rear panel. For metric
system installations, 6 mm adapters are supplied with each instrument to be used
if needed. The sample or calibration gas flowing through the system is monitored by a flowmeter downstream from the sensor.
The gases delivered to the instrument should be at constant pressures and
flow rates and must exit freely into the ambient atmosphere. The Span, Zero
and Sample gases should be delivered at constant pressures of about 10 psig
(Range 5-20). The flow rate must be maintained at about 700 cc/minute (Range
600-1000) and must exit freely into atmospheric pressure.
Figure 2-4 is the flow diagram for the sampling system. In the standard
instrument, calibration gases (zero and span) can be connected directly to the
Sample In port by teeing to the port with appropriate valves. The shaded portions of the diagram show the components added when the –C and/or –F
options are ordered. The solenoid valves, when supplied, are installed inside the
3020M enclosure and are regulated by the instruments internal electronics. The
flame arrestors, when supplied, are installed in the Gas Connector Panel.
Span In
Supplied at
constant
pressure
of 10 psig
Components in the shaded area are in
the -C option (internal control valves)
only and are not shown in the piping
diagram above.
Zero In
Sample In
Sensor
Solenoid
Valves
Needle Valves
to adjust the flow
Flowmeter
(Maintain at 700 cc/minute)
Exhaust Out
into athmosphere
Figure 2-4: Flow Diagram
Teledyne Analytical Instruments
2-5
2 Operational Theory
2.5
Model 3020M
Electronics and Signal Processing
The Model 3020M Percent Oxygen Analyzer uses an 8031 microcontroller with 32 kB of RAM and 128 kB of ROM to control all signal processing, input/output, and display functions for the analyzer. System power
is supplied from a universal power supply module designed to be compatible
with most international power sources. See chapter 5 Maintenance for the
location of the power supply and the main electronic PC boards. The Paramagnetic Sensor has its own dedicated universal 24V power supply that
feeds the sensors internal heater and electronics.
The signal processing electronics including the microprocessor, analog
to digital, and digital to analog converters are located on the main PCB, on
the front door (see figure 5-1). The preamplifier board is mounted on top of
the motherboard. These boards are accessible by opening the front door of
the explosion proof enclosure. Figure 2-5 is a block diagram of the Analyzer
electronics.
In the presence of oxygen the chamber is rotated in the magnetic field,
but the cell generates a current to countertorque the oxygen action. This
current is converted to a voltage, which is preamplified in the sensor internal
electronics.
The preamplified signal (0-1 Volt) is fed to the 3020M amplifier for
minor processing.
The digital concentration signal along with input from the control panel
is processed by the microprocessor, and appropriate control signals are
directed to the display, alarms and communications port. The same digital
information is also sent to a 12 bit digital to analog converter that produces
the 4-20 mA dc and the 0-1 V dc analog concentration signal outputs, and
the analog range ID outputs.
Signals from the power supply are also monitored, and through the
microprocessor, the system failure alarm is activated if a malfunction is
detected.
2.6 Temperature Control
For accurate analysis the sensor temperature is controlled internally at
55 C.
O
2-6
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Operational Theory 2
Sensor Amplifier
Sensor
A to D
Converter
24V Power
Supply
Auto
Range
Power
Supply
System
Failure
Alarm
MicroProcessor
Displays
Self Test
Signal
0-1 V
0-1 V
Concentration
D to A
Converter
Range
Figure 2-5: Block Diagram of the Model 3020M Electronics
Teledyne Analytical Instruments
2-7
2 Operational Theory
2-8
Teledyne Analytical Instruments
Model 3020M
Percent Paramagnetic Oxygen Analyzer
Installation 3
Installation
Installation of the Model 3020M Analyzer includes:
1. Unpacking
2. Mounting
3. Gas connections
4. Electrical connections
5. Testing the system.
3.1
Unpacking the Analyzer
The analyzer is shipped with all the materials you need to install and
prepare the system for operation. Carefully unpack the analyzer and inspect
it for damage. Immediately report any damage to the shipping agent.
3.2
Mounting the Analyzer
The Model 3020M is designed for bulkhead mounting in hazardous
environments. There are four mounting lugs—one in each corner of the
enclosure. The outline drawing, at the back of this manual, gives the mounting hole size and spacing. The drawing also contains the overall dimensions.
Do not forget to allow an extra 13/8" for the hinges.
Be sure to allow enough space in front of the enclosure to swing the
door open—a 16 1/4" radius, as shown in Figure 3-2.
All electrical connections are made via cables which enter the explosion-proof housing through ports in its side. No conduit fittings are supplied.
The installer must provide two 3/4" NPT and two 1" NPT adapters and the
appropriate sealing conduit.
Teledyne Analytical Instruments
3-1
3 Installation
Model 3020M
CMSC
Figure 3-1: Front View of the Model 3020M (Simplified)
3-2
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Installation 3
Figure 3-2: Required Front Door Clearance
3.3
Electrical Connections
Figure 3-3 shows the Model 3020M Electrical Connector Panel. There
are terminal blocks for connecting power, communications, and both digital
and analog concentration outputs.
100-240 V
50/60Hz
3.0A Max.
Figure 3-3: Electrical Connector Panel
Teledyne Analytical Instruments
3-3
3 Installation
Model 3020M
For safe connections, ensure that no uninsulated wire extends outside of
the connectors they are attached to. Stripped wire ends must insert completely into terminal blocks. No uninsulated wiring should be able to come in
contact with fingers, tools or clothing during normal operation.
3.3.1 Primary Input Power
The universal power supply requires a 100-240 VAC, 50 or 60 Hz
power source. See Figure 3-4 for detailed connections.
DANGER: Power is applied to the instrument's circuitry as
long as the instrument is connected to the power
source. The Standby function switches power on or
off to the displays and outputs only.
Figure 3-4: Primary Input Power Connections
3.3.2 Fuse Installation
The fuse holders accept 5 x 20 mm, 4.0 A, T type (slow blow) fuses.
Fuses are not installed at the factory. Be sure to install the proper fuse as part
of installation. (See Fuse Replacement in chapter 5, maintenance.)
3.3.3 Analog Outputs
There are eight DC output signal connectors on the ANALOG OUTPUTS connector block. There are two connectors per output with the polarity noted. See Figure 3-5.
3-4
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Installation 3
Figure 3-5: Analog Output Connections
The outputs are:
0–1 VDC % of Range: Voltage rises linearly with increasing oxygen, from
0 V at 0 % to 1 V at full scale. (Full scale = 100%
of programmable range.)
0–1 VDC Range ID:
0.25 V = Low Range, 0.5 V = Medium Range,
0.75 V = High Range, 1 V = Air Cal Range.
4–20 mADC % Range: (-M Option) Current increases linearly with increasing oxygen, from 4 mA at 0 % to 20 mA at full
scale. (Full scale = 100% of programmable range.)
4–20 mADC Range ID: (-M Option) 8 mA = Low Range, 12 mA = Medium Range, 16 mA = High Range, 20 mA = Air
Cal Range.
Examples:
The analog output signal has a voltage which depends on the oxygen
concentration AND the currently activated analysis range. To relate the
signal output to the actual concentration, it is necessary to know what range
the instrument is currently on, especially when the analyzer is in the
autoranging mode.
The signal output for concentration is linear over the currently selected
analysis range. For example, if the analyzer is set on a range that was defined
as 0-10 % O2, then the output would be as shown in Table 3-1.
Teledyne Analytical Instruments
3-5
3 Installation
Model 3020M
Table 3-1: Analog Concentration Output—Example
Voltage Signal
Output (V dc)
% O2
0
1
2
3
4
5
6
7
8
9
10
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
Current Signal
Output (mA dc)
4.0
5.6
7.2
8.8
10.4
12.0
13.6
15.2
16.8
18.4
20.0
To provide an indication of the range, a second pair of analog output
terminals are used. They generate a steady preset voltage (or current when
using the current outputs) to represent a particular range. Table 3-2 gives the
range ID output for each analysis range.
Table 3-2: Analog Range ID Output—Example
Range
LO
Voltage (V)
0.25
Current (mA)
8
MED
0.50
12
HI
0.75
16
1.00
20
0-25% Non Programmable
NOTE:
Impendance across 4-20mA output should not exceed 1000
ohms. Note polarity of terminals when wiring to the analog
outputs. Do not short 0-1VDC output for long period of time,
or damage may occur to analyzer.
CAUTION: Voltage may be present on alarm contacts when line power
is disconnected.
3-6
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Installation 3
3.3.4 Alarm Relays
There are three alarm-circuit connectors on the alarm relays block
(under RELAY OUTPUTS) for making connections to internal alarm relay
contacts. Each provides a set of Form C contacts for each type of alarm.
Each has both normally open and normally closed contact connections. The
contact connections are indicated by diagrams on the connector panel. They
are capable of switching up to 3 amperes at 250 VAC into a resistive load.
See Figure 3-6.
Figure 3-6: Types of Relay Contacts
The connectors are:
Threshold Alarm 1:
• Can be configured as high (actuates when concentration is above threshold), or low (actuates when
concentration is below threshold).
• Can be configured as failsafe or non-failsafe.
• Can be configured as latching or non-latching.
• Can be configured out (defeated).
Threshold Alarm 2:
• Can be configured as high (actuates when concentration is above threshold), or low (actuates when
concentration is below threshold).
• Can be configured as failsafe or non-failsafe.
• Can be configured as latching or non-latching.
• Can be configured out (defeated).
System Alarm:
Actuates when DC power supplied to circuits is
unacceptable in one or more parameters. Permanently
configured as failsafe and latching. Cannot be defeated. Actuates if self test fails.
To reset a System Alarm during installation, disconnect power to the instrument and then reconnect it.
Further detail can be found in chapter 4, section 4-10.
Teledyne Analytical Instruments
3-7
3 Installation
Model 3020M
NOTE: Power may be present on this contacts even when power to
analyzer has been removed since this contacts are wired by
the customer.
3.3.5 Digital Remote Cal Inputs
Remote Zero and Span Inputs: The REMOTE SPAN and REMOTE ZERO inputs are on the DIGITAL INPUT terminal block. They
accept 0 V (OFF) or 24 V dc (ON) for remote control of calibration. (See
Remote Calibration Protocol below.)
ZERO:
Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the Zero mode. Either side may be
grounded at the source of the signal. 0 to 1 volt across the
terminals allows Zero mode to terminate when done. A
synchronous signal must open and close the external zero
valve appropriately. See 3.3.9 Remote Sensor and Solenoid
Valves. (With the –C option, the internal valves automatically
operate synchronously.)
SPAN:
Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the Span mode. Either side may be
grounded at the source of the signal. 0 to 1 volt across the
terminals allows Span mode to terminate when done. A
synchronous signal must open and close the external span
valve appropriately. See 3.3.9 Remote Sensor and Solenoid
Valves. (With the –C option, the internal valves automatically
operate synchronously.)
Cal Contact: This relay contact is closed while analyzer is spanning
and/or zeroing. (See Remote Calibration Protocol below.)
Remote Calibration Protocol: To properly time the Digital Remote
Cal Inputs to the Model 3020M Analyzer, the customer's controller must
monitor the CAL CONTACT relay.
When the contact is OPEN, the analyzer is analyzing, the Remote Cal
Inputs are being polled, and a zero or span command can be sent.
When the contact is CLOSED, the analyzer is already calibrating. It
will ignore your request to calibrate, and it will not remember that request.
Once a zero or span command is sent, and acknowledged (contact
closes), release it. If the command is continued until after the zero or span is
complete, the calibration will repeat and the Cal Relay Contact (CRC) will
close again.
For example:
3-8
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Installation 3
1) Test the CRC. When the CRC is open, Send a zero command
until the CRC closes (The CRC will quickly close.)
2) When the CRC closes, remove the zero command.
3) When CRC opens again, send a span command until the CRC
closes. (The CRC will quickly close.)
4) When the CRC closes, remove the span command.
When CRC opens again, zero and span are done, and the sample is
being analyzed.
Note: The remote probe connections (paragraph 3.3.9) provides
signals to ensure that the zero and span gas valves will be
controlled synchronously. If you have the –C Internal valve
option—which includes additional zero and span gas inputs—
the 3020M automatically regulates the zero, span and sample
gas flow.
3.3.6 Range ID Relays
There are four dedicated RANGE ID CONTACT relays. The first
three ranges are assigned to relays in ascending order—Low range is assigned to RANGE 1 ID, Medium range is assigned to RANGE 2 ID, and
High range is assigned to RANGE 3 ID. RANGE 4 ID is reserved for the
Air Cal Range (25%).
3.3.7 Network I/O
A serial digital input/output for local network protocol. At this printing,
this port is not yet functional. It is to be used in future versions of the instrument.
3.3.8 RS-232 Port
The digital signal output is a standard RS-232 serial communications
port used to connect the analyzer to a computer, terminal, or other digital
device. The pinouts are listed in Table 3-3.
Table 3-3: RS-232 Signals
RS-232 Sig
DCD
RD
TD
RS-232 Pin
1
2
3
Purpose
Data Carrier Detect
Received Data
Transmitted Data
Teledyne Analytical Instruments
3-9
3 Installation
Model 3020M
DTR
COM
DSR
RTS
CTS
RI
4
5
6
7
8
9
Data Terminal Ready
Common
Data Set Ready
Request to Send
Clear to Send
Ring Indicator
The data sent is status information, in digital form, updated every two
seconds. Status is reported in the following order:
• The concentration in percent
• The range in use (HI, MED, LO)
• The span of the range (0-100 %, etc)
• Which alarms—if any—are disabled (AL–x DISABLED)
•
Which alarms—if any—are tripped (AL–x ON).
Each status output is followed by a carriage return and line feed.
Four input functions using RS-232 have been implemented to date.
They are described in Table 3-4.
Table 3-4: Commands via RS-232 Input
Command
Description
as
Immediately starts an autospan.
az
Immediately starts an autozero.
st
Toggling input. Stops/Starts any status message output from
the RS-232, until st is sent again.
The RS-232 protocol allows some flexibility in its implementation.
Table 3-5 lists certain RS-232 values that are required by the 3020M implementation.
Table 3-5: Required RS-232 Options
Parameter
Baud
Byte
Parity
Stop Bits
Message Interval
3-10
Setting
2400
8 bits
none
1
2 seconds
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Installation 3
3.3.9 Remote Sensor and Solenoid Valves
The 3020M is a single-chassis instrument, which has its own sensor
and, in the –C option, its own gas-control solenoid valves. The REMOTE
SENSOR connector is not used, and the SOLENOID RETURN connectors
are used (without the –C option) to synchronize external gas control valves.
See Figure 3-7.
16-2
15-2
14-2
13-2
16-1
15-1
14-1
13-1
+15 VDC
+15 VDC
+15 VDC
+15 VDC
SAMPLE (return)
SPAN (return)
ZERO (return)
EXHAUST (return)
Matching
Circuitry
(If
Necessary)
Remote Solenoid 1
Remote Solenoid 2
Remote Solenoid 3
Remote Solenoid 4
Remote Solenoid 1
Remote Solenoid 2
Remote Solenoid 3
Remote Solenoid 4
Figure 3-7: Remote Solenoid Return Connector Pinouts
The voltage from these outputs is nominally 0 V for the OFF and
15 V dc for the ON conditions. The maximum combined current that can be
pulled from these output lines is 100 mA. (If two lines are ON at the same
time, each must be limited to 50 mA, etc.) If more current and/or a different
voltage is required, use a relay, power amplifier, or other matching circuitry
to provide the actual driving current.
In addition, each individual line has a series FET with a nominal ON
resistance of 5 ohms (9 ohms worst case). This could limit the obtainable
voltage, depending on the load impedance applied. See Figure 3-8.
Figure 3-8: FET Series Resistance
Teledyne Analytical Instruments
3-11
3 Installation
3.4
Model 3020M
Gas Connections
Figure 3-9 is an illustration of the Gas Connector Panel. Optional gas
connections are shown in shaded blocks.
The unit is manufactured with 1/4 inch tube fittings only. Adapters must
be used for metric tubing. (At least 6 mm is recommended.)
For a safe connection:
1. Insert the tube into the tube fitting, and finger-tighten the nut until
the tubing cannot be rotated freely, by hand, in the fitting. (This
may require an additional 1/8 turn beyond finger-tight.)
2. Hold the fitting body steady with a backup wrench, and with
another wrench rotate the nut another 11/4 turns.
SCCM
Figure 3-9: Gas Connector Panel
SAMPLE IN: In the standard model, gas connections are made at the
SAMPLE IN and EXHAUST OUT connections. Calibration gases must be
Tee'd into the Sample inlet with appropriate valves.
3-12
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Installation 3
The gas pressure in should be regulated at about 10 psig. Pressures
between 5 and 20 psig are acceptable as long as the pressure, once established, will keep the front panel flowmeter reading in an acceptable range of
700 cc/minute.
If greater flow is required for improved response time, install a bypass
in the sampling system upstream of the analyzer input.
ZERO IN and SPAN IN: These are additional ports, included on
models with the –C option, for inputting span gas and zero gas. There are
electrically operated valves inside for automatic switching between sample
and calibration gases. These valves are completely under control of the
3020M Electronics. They can be externally controlled only indirectly
through the Remote Cal Inputs, described below.
Pressure, flow, and safety considerations are the same as prescribed for
the SAMPLE IN inlet, above.
EXHAUST OUT: Exhaust connections must be consistent with the
hazard level of the constituent gases. Check Local, State, and Federal laws,
and ensure that the exhaust stream vents to an appropriately controlled area if
required.
3.5
Testing the System
Before plugging the instrument into the power source:
• Check the integrity and accuracy of the gas connections. Make
sure there are no leaks.
• Check the integrity and accuracy of the electrical connections.
Make sure there are no exposed conductors
• Check that sample pressure is between 5 and 20 psig, according
to the requirements of your process, and the exhaut line is open to
atmosphere.
Power up the system, and test it by performing the following
operations:
1. Repeat the Self-Diagnostic Test as described in chapter 5, section
5.5.
Teledyne Analytical Instruments
3-13
3 Installation
3-14
Model 3020M
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Operation 4
Operation
4.1
Introduction
Once the analyzer has been installed, it can be configured for your
application. To do this you will:
•
•
•
•
•
Establish and start an automatic calibration cycle, if desired.
(Electrically operated valves required.)
Define the three user selectable analysis ranges. Then choose
autoranging or select a fixed range of analysis, as required.
Calibrate the instrument.
Set alarm setpoints, and modes of alarm operation (latching,
failsafe, etc).
Establish a security password, if desired, requiring Operator to
log in.
Before you configure your 3020M these default values are in effect:
Ranges: LO = 0-5 %, MED = 0-25 %, HI = 0-100 %.
Auto Ranging: ON
Alarm Relays: Defeated, 10 %, HI, Not failsafe, Not latching.
Zero: Auto, every 0 days at 0 hours.
Span: Auto, at 20.9 %, every 0 days at 0 hours.
Password: T E T A I
4.2
Using the Controls
To get the proper response from these controls, turn the control toward
the desired action (ESCAPE or ENTER—DOWN or UP), and then release it.
Turn-and-release once for each action. For example, turn-and-release twice
toward UP to move the VFD screen two selections upwards on the list of
options (menu).
Teledyne Analytical Instruments
4-1
4 Operation
Model 3020M
The item that is between arrows on the screen is the item that is currently selectable by choosing ENTER (turn-and-release toward ENTER with
the ESCAPE/ENTER control).
In these instructions, to ENTER means to turn-and-release toward
ENTER, and To ESCAPE means to turn-and-release towards ESCAPE. To
scroll UP (or scroll DOWN) means to turn-and-release toward UP (or
DOWN) as many times as necessary to reach the required menu item.
4.2.1 Mode/Function Selection
When the analyzer is first powered up, and has completed its initialization and self diagnostics, ESCAPE toggles the instrument between the
ANALYZE screen (Analysis Mode) and the MAIN MENU screen (Setup
Mode). The ANALYZE screen is the only screen of the Analysis Mode.
The MAIN MENU screen is the top level in a series of screens used in
the Setup Mode to configure the analyzer for the specific application. The
DOWN/UP commands scroll through the options displayed on the VFD
screen. The selectable option appears between arrows. When you reach the
desired option by scrolling, ENTER the selection as described below.
ESCAPE takes you back up the hierarchy of screens until you reach the
MAIN MENU again. ESCAPING any further just toggles between the
MAIN MENU and the ANALYZE screen.
4.2.1.1 Analysis Mode
This is the normal operating mode. The analyzer monitors the oxygen
content of the sample, displays the percent of oxygen, and warns of any
alarm conditions. Either control switches you to Setup Mode. Setup Mode
switches back to Analyze Mode if no controls are used for more than
five seconds.
4.2.1.2 Setup Mode
The MAIN MENU consists of 12 functions you can use to customize
and check the operation of the analyzer. Figure 4-1 shows the functions
available with the 3020M. They are listed here with brief descriptions:
1 AUTO-CAL: Used to define and/or start an automatic calibration
sequence.
2 PWD: Used to establish password protection or change the
existing password.
3 LOGOUT: Logging out prevents unauthorized tampering with
the analyzer settings.
4-2
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Operation 4
Setup Mode
AUTO-CAL
Analysis Mode
PSWD
ANALYZE
LOGOUT
Perform Oxygen
Analysis on
Sample
VERSION
SELF-TEST
Contrast Function is DISABLED
Auto-Cal On/Off
Span/Zero Timing
(Refer to Section 1.6)
SPAN
Auto-Cal Start
ZERO
Establish or
ALARMS
Change a
Password
RANGE
Log Out and
CONTRAST
Enable Security
Set Instrument
Zero
STANDBY
Display SoftSENSOR
Set Alarm
ware Version
Setpoints
Set Alarm
Modes of
Operation
Perform Self
Diagnostic
Routine
Set Instrument
Span
Set Display
Contrast
Define Analysis
Ranges
(Not Used)
ON, but No
Displays or Outputs
Figure 4-1: Modes and Functions
4
VERSION: Displays Manufacturer, Model, and Software version
of the instrument.
5
SELF-TEST: The instrument performs a self-diagnostic routine
to check the integrity of the power supply, output boards and
amplifiers.
SPAN: Set up and/or start a span calibration.
ZERO: start a zero calibration.
ALARMS: Used to set the alarm setpoints and determine
whether each alarm will be active or defeated, HI or LO acting,
latching or not, and failsafe or not.
RANGE: Used to set up three analysis ranges that can be
switched automatically with auto-ranging or used as individual
fixed ranges.
6
7
8
9
Teledyne Analytical Instruments
4-3
4 Operation
Model 3020M
10 STANDBY: Remove power to outputs and displays, but
maintain power to internal circuitry.
11 Sensor (Not Used).
Any function can be selected at any time. Just scroll through the MAIN
MENU with the DOWN/UP control to the appropriate function, and ENTER
it. The analyzer will immediately start that function, unless password
restrictions have been assigned. (Password assignment is explained further
on.)
All of these functions are described in greater detail in the procedures
starting in section 4.3. The VFD screen texts used to illustrate the procedures
are reproduced in a Monospaced type style.
4.2.2 Data Entry
4.2.2.1 ENTER
When the selected option is a function on the Main Menu screen, the
function name appears between the arrows on the screen. You activate the
function by turning the ESCAPE/ENTER control to ENTER.
When the selected option is a function or subfunction, ENTER moves
the display to the VFD screen for that function or subfunction.
When the selected option is a modifiable item, the DOWN/UP control
can be used to increment or decrement that modifiable item to the value or
action you want. Then you ENTER the item, which also puts you into the
next field to continue programming.
When the last field is entered, ENTER takes you to the next screen in the
process, or if the process is completed, ENTER takes you back to the ANALYZE screen.
4.2.2.2 ESCAPE
A turn-and-release toward ESCAPE moves the blinking to the next
field on the left. When you are on the leftmost field, another ESCAPE takes
you back to the previous screen.
If you do not wish to continue a function, you can abort the session by
escaping to the leftmost field, and then issuing another ESCAPE. Escaping a
function takes the analyzer back to the previous screen, or to the ANALYZE
Function, depending on the function escaped.
4-4
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
4.3
Operation 4
The AUTO-CAL Function
When proper automatic valving is connected (see chapter 3, installation), the Analyzer can cycle itself through a sequence of steps that automatically zero and span the instrument.
Note: If you require highly accurate timing of your AUTO-CAL, use
external AUTO-CAL control where possible. The internal clock
in the Model 3020M is accurate to 2-3 %. Accordingly, internally scheduled calibrations can vary 2-3 % per day.
To setup an AUTO-CAL cycle:
Scroll to AUTO-CAL, and ENTER. A new screen for Span/Zero set
appears.
Span OFF Nxt: 0d 0h
Zero OFF Nxt: 0d 0h
If SPAN (or ZERO) is not flashing, scroll with DOWN/UP control to
SPAN (or ZERO), then ENTER. (You won’t be able to set OFF to ON if a
zero interval is entered.) A Span Every ... (or Zero Every ...) screen appears.
Span Every 0 d
Start 0 h from now
Use DOWN/UP control to set an interval value, and ENTER. Then use
DOWN/UP to set a start-time value, and ENTER.
Span OFF Nxt: 0d 0h
Zero OFF Nxt: 0d 0h
When the Span/Zero values screen appears, use the scroll DOWN to
blink the OFF/ON field of the SPAN (or ZERO) function. Use DOWN/UP
to set the OFF/ON field to ON. You can now turn these fields ON because
there is a nonzero span interval defined.
4.4
The PWD (Password) Function
Security can be established by choosing a 5 digit password from the
standard ASCII character set. If you decide NOT to employ password
security, use the default password TETAI. This password will be displayed
automatically by the microprocessor. The operator just ENTERs it to be
allowed total access to the instrument’s features.
Once a unique password is assigned and activated, the operator MUST
enter the UNIQUE password to gain access to any of the set-up functions
(except to enter the password). However, the instrument will continue to
Teledyne Analytical Instruments
4-5
4 Operation
Model 3020M
analyze sample gas and report on alarm conditions without entering the
password.
• Only one password can be defined.
• After a password is assigned, the operator must log out to
activate it. Until then, anyone can continue to operate the
instrument without entering the new password.
• To defeat the security after a unique password is activated, the
password must be changed back to TETAI.
NOTE: If you use password security, it is advisable to keep a copy of
the password in a separate, safe location.
4.4.1 Entering the Password
To install a new password or change a previously installed password,
you must key in and enter the old password first. If the default password is in
effect, issuing the ENTER command will enter the default TETAI password
for you.
Scroll to PWD, and ENTER to select the password function. Either the
TETAI default password or AAAAA place-holder password for an existing
password will appear on screen depending on whether or not a password has
been previously installed.
TETAI
Enter PWD
or
AAAAA
Enter PWD
The screen prompts you to enter the current password. If you are not
using password protection, ENTER to accept TETAI as the default password.
If a password has been previously installed, enter the password using ENTER to scroll through the letters, and the DOWN/UP keys to change the
letters to the proper password. The last ENTER enters the password.
If the password is accepted, the screen will indicate that the password
restrictions have been removed and you have clearance to proceed.
PWD Restrictions
Removed
In a few seconds, if you do not ESCAPE, you will be given the opportunity to change this password or keep it and go on.
4-6
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Operation 4
Change Password?
=Yes =No
ESCAPE to move on, or proceed as in Changing the Password, below.
4.4.2 Installing or Changing the Password
If you want to change the existing password: Proceed as above in
Entering the Password, until you are given the opportunity to change the
password:
Change Password?
=Yes =No
ENTER to change the password (to change either the default TETAI or
the previously assigned password), or ESCAPE to keep the existing password and move on.
If you choose ENTER to change the password, the password assignment screen appears.
TETAI
To Proceed
or
AAAAA
To Proceed
Enter the password using ENTER to scroll through the existing password letters, and DOWN/UP to change the letters to the new password. The
full set of 94 characters available for password use are shown in the table
below.
Characters Available for Password Definition:
A
K
U
_
i
s
}
)
3
=
B
L
V
`
j
t
→
*
4
>
C
M
W
a
k
u
!
+
5
?
D
N
X
b
l
v
"
'
6
@
E
O
Y
c
m
w
#
7
F
P
Z
d
n
x
$
.
8
G
Q
[
e
o
y
%
/
9
H
R
¥
f
p
z
&
0
:
I
S
]
g
q
{
'
1
;
J
T
^
h
r
|
(
2
<
When you have finished typing the new password, the last ENTER
enters it. A verification screen appears. The screen will prompt you to retype
your password for verification.
Teledyne Analytical Instruments
4-7
4 Operation
Model 3020M
AAAAA
Retype PWD To Verify
Wait a moment for the entry () screen. You will be given
clearance to proceed.
AAAAA
TO Proceed
ENTER the letters of your new password. Your password will be stored
in the microprocessor and the system will immediately switch to the ANALYZE screen, and you now have access to all instrument functions.
If all alarms are defeated, the ANALYZE screen appears as:
0.0
% Anlz
Range: 0 — 100
If an alarm is tripped, the second line will change to show which alarm
it is:
0.0
% Anlz
AL—1
NOTE:If you log off the system using the LOGOUT function in the
MAIN MENU, you will now be required to re-enter the password to gain access to any of the functions except password
(PWD).
4.5
The LOGOUT Function
By entering LOGOUT, you effectively log off the instrument, leaving
the system protected against tampering until the password is reentered. To
log out, scroll to place the LOGOUT function between the arrows, and
ENTER to log out. The screen will display the message:
Protected Until
Password Reentered
4.6
The VERSION Screen
Scroll through the MAIN MENU to VERSION, and ENTER. The
screen displays the manufacturer, model, and software version information.
4-8
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
4.7
Operation 4
The SELF-TEST Function
The Model 3020M has a built-in self-testing diagnostic routine. Preprogrammed signals are sent through the power supply, output board and sensor
circuit. The return signal is analyzed, and at the end of the test the status of
each function is displayed on the screen, either as OK or as a number between 1 and 3. (See System Self Diagnostic Test in chapter 5 for number
code.)
The self diagnostics are run automatically by the analyzer whenever the
instrument is turned on, but the test can also be run by the operator at will.
To initiate SELF-TEST during operation:
Use the DOWN/UP control to scroll through the MAIN MENU to
SELF-TEST. The screen will follow the running of the diagnostic.
RUNNING DIAGNOSTIC
Testing Preamp — 83
During preamp testing there is a countdown in the lower right corner of
the screen. When the testing is complete, the results are displayed.
Power: OK Analog: OK
Preamp: 3
The module is functioning properly if it is followed by OK. A number
indicates a problem in a specific area of the instrument. Refer to Chapter 5
Maintenance for number-code information. The results screen alternates for a
time with:
Press Any Key
To Continue...
Then the analyzer returns to the ANALYZE screen.
4.8
The SPAN Functions
The analyzer is calibrated using span gas.
For optimum accuracy, we recommend the analyzer be spanned in the
high range with span gas of a known oxygen concentration in the range of
70-90% of full scale.
Connect the calibration gases to the analyzer according to the instructions given in section 3.5, Gas Connections, observing all the prescribed
precautions.
Shut off the gas pressure before connecting it to the analyzer, and
be sure to control the pressure to about 10 psig when turning it back
on.
Teledyne Analytical Instruments
4-9
4 Operation
Model 3020M
Readjust the gas pressure into the analyzer until the flowrate (as read on
the analyzer’s SCCM flowmeter) settles at 700 cc/minute.
If you are using password protection, you will need to enter your
password to gain access to either of these functions. Follow the instructions
in section 4.4 to enter your password. Once you have gained clearance to
proceed, you can ENTER the ZERO or SPAN function.
4.8.1 Span Cal
SPAN is used to span calibrate the analyzer. Span calibration can be
performed using the automatic mode, where an internal algorithm compares
consecutive readings from the sensor to determine when the output matches
the span gas concentration. Span calibration can also be performed in
manual mode, where the operator determines when the span concentration
reading is acceptable and manually exits the function.
4.8.1.1
Auto Mode Spanning
Scroll to SPAN, and ENTER the SPAN function. The screen that
appears allows you to select whether the span calibration is to be performed
automatically or manually. Use the DOWN/UP control to toggle between
AUTO and MAN span settling. Stop when AUTO appears on the display.
Span: Settling: AUTO
For Next
Use ENTER to move to the next screen.
Span Val: 20.90
Span Mod #
Use DOWN/UP to start changing the oxygen-concentration. Use
ESCAPE/ENTER to blink the digit you are going to modify. Use DOWN/UP
again to change the value of the selected digit. When you have finished
typing in the concentration of the span gas you are using, repeatedly select
ENTER until the rightmost digit is reached, then, the next ENTER will exit the
Span Val screen. One more ENTER will enter the new span value, bring up
the next screen, and start the span calibration.
####
% Span
Slope=####
ppm/s
The beginning span value is shown in the upper left corner of the
display. As the span reading settles, the screen displays and updates information on Slope. Spanning automatically ends when the span output corresponds, within tolerance, to the value of the span gas concentration. Then the
instrument automatically returns to the ANALYZE mode.
4-10
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
4.8.1.2
Operation 4
Manual Mode Spanning
ENTER SPAN from the MAIN MENU to start the SPAN function.
The screen that appears allows you to select whether the span calibration is
to be performed automatically or manually.
Span: Settling:MAN
For Next
Use DOWN/UP to toggle between AUTO and MAN span settling. Stop
when MAN appears on the display. Use ENTER to move to the next screen.
Span Val: 20.90
Span Mod #
Use UP to permit modification (Mod #) of span value.
Use ESCAPE/ENTER to choose the digit, and use DOWN/UP to
choose the value of the digit.
When you have finished typing in the concentration of the span gas you
are using, repeatedly select ENTER until the rightmost digit is reached, then, the
next ENTER will exit the Span Val field. One more ENTER will enter the new
span value, bring up the next screen, and start the span calibration.
Once the span has begun, the microprocessor samples the output at a
predetermined rate. It calculates the difference between successive samplings
and displays this difference as Slope on the screen. It takes several seconds
for the first Slope value to display. Slope indicates rate of change of the Span
reading. It is a sensitive indicator of stability.
####
% Span
Slope=#### ppm/s
When the Span value displayed on the screen is sufficiently stable,
ENTER it. (Generally, when the Span reading changes by 1 % or less of the
full scale of the range being calibrated for a period of ten minutes it is sufficiently stable.) Once you ENTER it, the Span reading changes to the correct
value. The instrument then automatically enters the ANALYZE function.
4.9
Zero Cal
The ZERO function on the MAIN MENU is used to enter the zero
calibration function. Zero calibration can be performed in either the automatic or manual mode. In the automatic mode, an internal algorithm compares consecutive readings from the sensor to determine when the output is
within the acceptable range for zero. In the manual mode, the operator
determines when the reading is within the acceptable range for zero.
Make sure the zero gas is connected to the instrument.
Teledyne Analytical Instruments
4-11
4 Operation
Model 3020M
If you have trouble zeroing, you may need to skip to section 4.8.1.3
Cell Failure.
Auto Mode Zeroing
Select ZERO to enter the ZERO function. The ZERO screen allows
you to select whether the zero calibration is to be performed automatically or
manually. Use the DOWN/UP control to toggle between AUTO and MAN
zero settling. Stop when AUTO appears on the display.
Zero: Settling: AUTO
To Begin
ENTER to begin zeroing.
####
% Zero
Slope=#### ppm/s
The beginning zero level is shown in the upper left corner of the display. As the zero reading settles, the screen displays and updates information
on Slope (unless the Slope starts within the acceptable zero range and does
not need to settle further).
Then, and whenever Slope is less than 0.08 for at least 3 minutes,
instead of Slope you will see a countdown, such as 1 Left, 0 Left. These are
steps in the zeroing process that the system must complete, AFTER settling,
before it can go back to ANALYZE.
####
% Zero
1 Left=### ppm/s
The zeroing process will automatically conclude when the output is
within the acceptable range for a good zero. Then the analyzer automatically
returns to the ANALYZE screen.
Manual Mode Zeroing
ENTER the ZERO function. The screen that appears allows you to
select between automatic or manual zero calibration. Use DOWN/UP to
toggle between AUTO and MAN zero settling. Stop when MAN appears on
the display.
Zero: Settling: Man
To Begin
ENTER to begin the zero calibration. After a few seconds the first of
five zeroing screens appears. The number in the upper left hand corner is the
first-stage zero offset. The microprocessor samples the output at a predetermined rate. It calculates the differences between successive samplings and
4-12
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Operation 4
displays the rate of change as Slope= a value in parts per million per second
(ppm/s).
####
% Zero
Slope=#### ppm/s
NOTE:It takes several seconds for the true Slope value to display.
Wait about 10 seconds. Then, wait until Slope is sufficiently
close to zero before using ENTER to finish zeroing.
Generally, you have a good zero when Slope is less than 0.05 ppm/s
for about 30 seconds. When Slope is close enough to zero, ENTER it. In a
few seconds, the screen will update.
Once zero settling completes, the information is stored in the
microprocessor, and the instrument automatically returns to the ANALYZE
screen.
4.10 The ALARMS Function
The Model 3020M is equipped with 2 fully adjustable concentration
alarms and a system failure alarm. Each alarm has a relay with a set of form
“C" contacts rated for 3 amperes resistive load at 250 V ac. See Figure in
Chapter 3, Installation and/or the Interconnection Diagram included at the
back of this manual for relay terminal connections.
The system failure alarm has a fixed configuration described in chapter
3 Installation.
The concentration alarms can be configured from the front panel as
either high or low alarms by the operator. The alarm modes can be set as
latching or non-latching, and either failsafe or non-failsafe, or, they can be
defeated altogether. The setpoints for the alarms are also established using
this function.
Decide how your alarms should be configured. The choice will depend
upon your process. Consider the following four points:
1. Which if any of the alarms are to be high alarms, and which if
any are to be low alarms?
Setting an alarm as HIGH triggers the alarm when the oxygen
concentration rises above the setpoint. Setting an alarm as LOW
triggers the alarm when the oxygen concentration falls below the
setpoint.
Decide whether you want the alarms to be set as:
• Both high (high and high-high) alarms, or
Teledyne Analytical Instruments
4-13
4 Operation
Model 3020M
• One high and one low alarm, or
• Both low (low and low-low) alarms.
2. Are either or both of the alarms to be configured as failsafe?
In failsafe mode, the alarm relay de-energizes in an alarm
condition. For non-failsafe operation, the relay is energized in an
alarm condition. You can set either or both of the concentration
alarms to operate in failsafe or non-failsafe mode.
3. Are either of the alarms to be latching?
In latching mode, once the alarm or alarms trigger, they will
remain in the alarm mode even if process conditions revert back
to non-alarm conditions. This mode requires an alarm to be
recognized before it can be reset. In the non-latching mode, the
alarm status will terminate when process conditions revert to nonalarm conditions.
4. Are either of the alarms to be defeated?
The defeat alarm mode is incorporated into the alarm circuit so
that maintenance can be performed under conditions which
would normally activate the alarms.
The defeat function can also be used to reset a latched alarm.
(See procedures, below.)
If you are using password protection, you will need to enter your
password to access the alarm functions. Follow the instructions in section 4.4
to enter your password. Once you have clearance to proceed, ENTER the
ALARM function.
AL—1
AL—2
Choose Alarm
Use the DOWN/UP control to blink your choice of alarm, AL-1 or
AL-2. Then ENTER to move to the next screen.
AL—1 1.00 % HI
Dft—N Fs—N Ltch—N
Five parameters can be changed on this screen.
• Value of the alarm setpoint: AL–1 #### % (oxygen)
• Out-of-range direction: HI or LO
• Defeated? (Yes/No): Dft–Y/N
• Failsafe? (Yes/No): Fs–Y/N
• Latching? (Yes/No): Ltch–Y/N.
• To define the setpoint, use ENTER to blink AL–1 ####, if not
already blinking. Then use the DOWN/UP control to change the
4-14
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
•
•
•
Operation 4
number. Holding the control on the DOWN or UP position, while
the number changes, speeds up the incrementing or
decrementing. (Remember, the setpoint units are always % O2.)
To set the other parameters use ENTER to blink the desired
parameter. Then use DOWN/UP to change the parameter.
Once the parameters for the alarm have been set, ENTER the
ALARM function again, and repeat this procedure for next alarm.
To reset a latched alarm:
Go to Dft– and then assert either DOWN two times or UP two
times. (Toggle it to Y and then back to N.)
–OR –
Go to Ltch– and then assert either DOWN two times or UP two
times. (Toggle it to N and back to Y.)
Note: In this screen, ENTER moves you through the fields and
ESCAPE takes you back to the previous screen.
4.11 The RANGE Function
The RANGE function allows the operator to program up to three
concentration ranges to correlate with the DC analog outputs. If no ranges
are defined by the user, the instrument defaults to:
Low = 0–5 %
Med = 0–25 %
High = 0–100%.
The Model 3020M is set at the factory to default to autoranging. In this
mode, the microprocessor automatically responds to concentration changes
by switching ranges for optimum readout sensitivity. If the current range
limits are exceeded, the instrument will automatically shift to the next higher
range. If the sample concentration falls below the full scale of the next lower
range, the instrument will switch to that range. A corresponding shift in the
DC percent-of-range output, and in the range ID outputs, will be noticed.
The autoranging feature can be overridden so that analog output stays
on a fixed range regardless of the oxygen concentration detected. If the
concentration exceeds the upper limit of the range, the DC output will
saturate at 1 VDC or 20 mA at the current output.
However, the digital readout and the RS-232 output of the concentration are unaffected by the fixed range. They continue to read accurately with
full precision. See Front Panel description in Chapter 1.
Teledyne Analytical Instruments
4-15
4 Operation
Model 3020M
The automatic air calibration range is always 0-25 % and is not programmable.
4.11.1 Setting the Analog Output Ranges
To set the ranges, ENTER the RANGE function mode by selecting
RANGE from the MAIN MENU. The RANGE screen appears.
L—5.00 M—25.00
H—100.00 Mode—AUTO
Use the DOWN/UP control to the range to be set: low (L), medium (M),
or high (H).
Use the DOWN/UP control to enter the upper value of the range (all
ranges begin at 0 %). Repeat for each range you want to set. ENTER to
accept the values and return to the Analysis Mode. (See note below.)
Note: The ranges must be increasing from low to high, for example,
if range 1 is set as 0–10 % and range 2 is set as 0–100 %, range
3 cannot be set as 0–25 % since it is lower than range 2.
4.11.2 Automatic Ranging
After defining your analysis ranges, set Mode– to AUTO, if not already
there. ESCAPE to re-enter the ANALYZE screen using the fixed range.
4.11.3 Fixed Range Analysis
The autoranging mode of the instrument can be overridden, forcing the
analyzer DC outputs to stay in a single predetermined range.
To switch from autoranging to fixed range analysis, ENTER the
RANGE function by selecting RANGE from the MAIN MENU.
Use the DOWN/UP control to move AUTO between the arrows.
Use the DOWN/UP control to switch from AUTO to FX/LO, FX/
MED, or FX/HI to set the instrument on the desired fixed range (low, medium, or high).
L—5.00 M—25.00
H—100.00 Mode—FX/LO
or
L—5.00 M—25.00
H—100.00 Mode—FX/MED
or
4-16
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Operation 4
L—5.00 M—25.00
H—100.00 Mode—FX/HI
ESCAPE to re-enter the ANALYZE screen using the fixed range.
Note: When performing analysis on a fixed range, if the oxygen
concentration rises above the upper limit, as established by
the setup, for that particular range, the output saturates at 1
V dc (20 mA at optional mA dc output). However, the digital
readout and the RS-232 output continue to read the true value
of the oxygen concentration regardless of the analog output
range.
Contrast Function is DISABLED
(Refer to Section 1.6)
4.12 The CONTRAST Function
If you cannot read anything on the display after first powering up:
1. Observe LED readout.
a. If LED meter reads all eights and dots, go to step 3.
b. If LED meter displays anything else, go to step 2.
2. Disconnect power to the Analyzer and reconnect again. LED
meter should now read all eights and dots.
4.13 The STANDBY Function
In STANDBY, the analyzer’s internal circuits are powered, but there
are no displays or outputs from the analyzer.
WARNING: THE POWER CABLE MUST BE UNPLUGGED TO
FULLY DISCONNECT POWER FROM THE INSTRUMENT. WHEN THE ACCESS DOOR IS OPEN AND
THE POWER CABLE IS CONNECTED, EXTRA
CARE IS REQUIRED TO AVOID CONTACT WITH
LIVE ELECTRICAL CIRCUITS.
CAUTION:
If you disconnect the primary power source from
the analyzer, then on re-energizing, you will be
required to choose to keep the configuration you
previously programmed into your instrument in the
Setup Mode, or it will reset to factory defaults.
Teledyne Analytical Instruments
4-17
4 Operation
Model 3020M
Use STANDBY whenever you want to power down without danger
loosing all of the configuration you programmed into your instrument in the
Setup Mode.
4.14 The Analysis Mode
This is the normal operating mode of the analyzer. In this mode the
analyzer is monitoring the sample, measuring and displaying the amount of
oxygen, and reporting alarm conditions.
Normally, the Version, Self-Test, Sensor, Main Menu, and the Span
and Zero (in auto-settling mode) functions automatically switch back to the
Analysis Mode ANALYZE screen when they have completed their assigned
operations. After four or five seconds in the MAIN MENU without any action
by the operator, the analyzer automatically switches itself back to the ANALYZE screen. ESCAPE, asserted one or more times, depending on the
starting point, also switches the analyzer back to the ANALYZE screen.
4.15 Sensor Function
Not Used on the 3020M. This function tests micro-fuel sensor output
on the 3020 for those that are using it.
4-18
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Maintenance 5
Maintenance
5.1
Routine Maintenance
Aside from normal cleaning and checking for leaks at the gas connections, routine maintenance is limited to replacing Paramagnetic Sensor and
fuses, and recalibration. For recalibration, see Section 4.4 The Zero and
Span Functions.
WARNING: SEE WARNINGS ON TITLE PAGE OF THIS MANUAL.
5.2
Major Internal Components
All internal components are accessed by unbolting and swinging open
the front cover, as described earlier. The major internal component locations
are shown in Figure 5-1, and the fuse receptacle is shown in Figure 5-2.
The 3020M contains the following major internal components:
• Paramagnetic Sensor
• Paramagnetic Sensor 24Vdc Power Supply
• Customer Interface PCB (Power Supply on bottom surface)
• Preamp PCB (Contains Microprocessor)
• Front Panel PCB (Contains Displays)
5 digit LED meter
2 line, 20 character, alphanumeric, VFD display
• Solenoid Operated Gas Control Valves (–C option only).
Teledyne Analytical Instruments
5-1
5 Maintenance
Model 3020M
24V Sensor Power Supply
located behind Customer
Interface Board.
Display PCB
Main PCB
Preamp PCB
Door Subassembly
(Cross Section)
Box Subassembly
(Gas Connector Panel
not shown)
Customer Interface Panel
(Electrical Connector Panel)
Gas Control Valves
Sensor
Figure 5-1: Major Internal Components
To swing open the cover panel, remove all screws.
5.3
Sensor Replacement
The Paramagnetic Sensor is maintenance free, and has a long life.
Should the sensor fail, contact the factory for service and replacement.
5.4
Fuse Replacement
The 3020M requires two 5 x 20 mm, 4 A, T type (Slow Blow) fuses.
The fuses are located inside the explosion proof housing on the Electrical
Connector Panel, as shown in Figure 5-2. To replace a fuse:
1. Disconnect the Unit from its power source.
2. Place a small screwdriver in the notch in the fuse holder cap,
push in, and rotate 1/4 turn. The cap will pop out a few
millimeters. Pull out the fuse holder cap and fuse, as shown in
Figure 5-3.
5-2
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Maintenance 5
Figure 5-2: Removing Fuse Cap and Fuse from Holder
2. Replace fuse by reversing process in step 1.
5.5
System Self Diagnostic Test
Use the DOWN/UP control to scroll through the MAIN MENU to
SELF-TEST. The screen will follow the running of the diagnostic.
RUNNING DIAGNOSTIC
Testing Preamp — 83
During preamp testing there is a countdown in the lower right corner of
the screen. When the testing is complete, the results are displayed.
Power: OK Analog: OK
Preamp: 3
The module is functioning properly if it is followed by OK. A number
indicates a problem in a specific area of the instrument. Refer to Table 5-1
for number-code information. The results screen alternates for a time with:
Press Any Key
To Continue...
The following failure codes apply:
Table 5-1: Self Test Failure Codes
Power
0
1
2
3
OK
5 V Failure
15 V Failure
Both Failed
Teledyne Analytical Instruments
5-3
5 Maintenance
Model 3020M
Analog
0
1
2
3
OK
DAC A (0–1 V Concentration)
DAC B (0–1 V Range ID)
Both Failed
Preamp
0
1
2
3
OK
Zero too high
Amplifier output doesn't match test input
Both Failed
Note that preamp test does not tell anything about sensor condition. If
preamp is ok but instrument is not performing then sensor, sensor power
supply, and sensor interconnection cable are suspects.
5.6
Output Goes Negative
If during operation of the instrument, gas pressure exceeds 20 psig, the
output of the sensor may reverse polarity, readings on the display may go
negative. If this happens:
Turn power off by unplugging the instrument.
Turn gas pressure to less than 20 psig.
Turn power on again.
Return instrument to analyze mode.
5-4
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Appendix
Appendix
A-1 Specifications
Packaging: Explosion-proof. Bulkhead mount.
Sensor: Paramagnetic Sensor.
Ranges: Three user definable ranges from 0-5 % to 0100 %.
Autoranging with range ID output.
Sample System: Flow indicator visible from front of unit.
Positive pressure service.
Vacuum service (optional).
Auto Cal / Auto Zero. (Available with optional, electrically operated valves.)
Alarms: One system-failure alarm contact to detect
power failure.
Two adjustable concentration threshold alarms
with fully programmable setpoints.
Diagnostics: Start-up or on-demand self-testing function.
Displays: 2 line by 20 alphanumeric, VFD screen, and
one 5 digit LED display.
Digital Interface: Full duplex RS-232 communications port.
Power: Universal power supply 100-240 V ac, at 50
or 60 Hz, 70 Watts max.
Operating Temperature: 0-45 °C
Accuracy: ±2% of full scale at constant temperature.
±5% of full scale over operating temperature
Teledyne Analytical Instruments
A-1
Appendix
Model 3020M
range, on factory default analysis ranges, once
thermal equilibrium has been achieved.
Analog outputs: 0-1 V dc percent-of-range
0-1 V dc range ID.
4-20 mADC percent-of-range
4-20 mADC range ID.
Password Access: Can be user-configured for password
protection.
A-2
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
Appendix
A-2 Recommended 2-Year Spare Parts List
QTY.
PART NUMBER
DESCRIPTION
1
C62371B
Display PCB
1
D70046
Customer Interface PCB
1
C70043
Percent Preamplifier Board
1*
C62365-C
Main PCB (NO 4-20 mA)
1*
C62365-A
Main PCB (4-20 mA)
4
F1295
Fuse, 4A, 250V, 5x20 mm, T (Slow Blow)
A minimum charge is applicable to spare parts orders.
Note: Orders for replacement parts should include the part number (if
available) and the model and serial number of the instrument for
which the parts are intended.
Orders should be sent to:
Teledyne Analytical Instruments
16830 Chestnut Street
City of Industry, CA 91749-1580
Phone (626) 961-9221, Fax (626) 961-2538
TWX (910) 584-1887 TDYANYL COID
or your local representative.
A-3 Drawing List
D-70000 Outline Drawing
D-69991 Wiring Diagram
D-69990 Piping Diagram
D-69989 Interconnection Diagram
Teledyne Analytical Instruments
A-3
Appendix
NOTE:
A-4
Model 3020M
The MSDS on this material is available upon request
through the Teledyne Environmental, Health and
Safety Coordinator. Contact at (626) 934-1592
Teledyne Analytical Instruments
Percent Paramagnetic Oxygen Analyzer
A-5
Appendix
Zero Cal
The ZERO function on the MAIN MENU is used to enter the zero
calibration function. Zero calibration can be performed in either the automatic or manual mode. In the automatic mode, an internal algorithm compares consecutive readings from the sensor to determine when the output is
within the acceptable range for zero. In the manual mode, the operator
determines when the reading is within the acceptable range for zero.
Make sure the zero gas is connected to the instrument.
If you have trouble zeroing, you may need to skip to section 4.8.1.3
Cell Failure.
Auto Mode Zeroing
Select ZERO to enter the ZERO function. The ZERO screen allows
you to select whether the zero calibration is to be performed automatically or
manually. Use the DOWN/UP control to toggle between AUTO and MAN
zero settling. Stop when AUTO appears on the display.
Zero: Settling: AUTO
To Begin
ENTER to begin zeroing.
####
% Zero
Slope=#### ppm/s
The beginning zero level is shown in the upper left corner of the display. As the zero reading settles, the screen displays and updates information
on Slope (unless the Slope starts within the acceptable zero range and does
not need to settle further).
Then, and whenever Slope is less than 0.08 for at least 3 minutes,
instead of Slope you will see a countdown, such as 1 Left, 0 Left. These are
steps in the zeroing process that the system must complete, AFTER settling,
before it can go back to ANALYZE.
####
% Zero
1 Left=### ppm/s
The zeroing process will automatically conclude when the output is
within the acceptable range for a good zero. Then the analyzer automatically
returns to the ANALYZE screen.
Teledyne Analytical Instruments
A-5
Appendix
Model 3020M
Manual Mode Zeroing
ENTER the ZERO function. The screen that appears allows you to
select between automatic or manual zero calibration. Use DOWN/UP to
toggle between AUTO and MAN zero settling. Stop when MAN appears on
the display.
Zero: Settling: Man
To Begin
ENTER to begin the zero calibration. After a few seconds the first of
five zeroing screens appears. The number in the upper left hand corner is the
first-stage zero offset. The microprocessor samples the output at a predetermined rate. It calculates the differences between successive samplings and
displays the rate of change as Slope= a value in parts per million per second
(ppm/s).
####
% Zero
Slope=#### ppm/s
NOTE:It takes several seconds for the true Slope value to display.
Wait about 10 seconds. Then, wait until Slope is sufficiently
close to zero before using ENTER to finish zeroing.
Generally, you have a good zero when Slope is less than 0.05 ppm/s
for about 30 seconds. When Slope is close enough to zero, ENTER it. In a
few seconds, the screen will update.
Once zero settling completes, the information is stored in the
microprocessor, and the instrument automatically returns to the ANALYZE
screen.
A-6
Teledyne Analytical Instruments