Yokogawa pH and ORP Learning Handbook (Reference)
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
www.yokogawa.com
pH and ORP Learning Handbook
©
TI12B00A20-01E
Copyright March 2014
1st edition
w w w.yo ko g a wa.c o m
Table of Contents
1
Introduction.................................................................................................................... 4
2
Basics on pH theory........................................................................................................4
2.1
Concept of pH..........................................................................................................................4
2.2
The pH scale............................................................................................................................ 5
2.3
Measuring the pH scale............................................................................................................ 6
2.4
Principle of potentiometric pH measurement................................................................................. 7
2.5
Hydrogen electrodes, the basic principle..................................................................................... 7
2.5.1
Composition of the glass electrode...................................................................................... 9
2.5.2
Composition of the reference electrode................................................................................ 9
2.5.3
The measuring circuit with a glass and a reference electrode............................................... 10
2.5.4
More information about the glass electrode.........................................................................11
2.5.5
More information about the reference system...................................................................... 12
2.5.6
Construction of the temperature Electrode........................................................................... 15
2.5.7
The combined pH sensors................................................................................................. 15
2.6
The effect of temperature ........................................................................................................ 16
2.7
Isolation resistance.................................................................................................................. 18
2.8
Buffer solutions....................................................................................................................... 18
2.9
Periodic maintenance and calibration of pH sensors................................................................... 19
2.10
Differential electrodes.............................................................................................................. 23
3 Basics on ORP...............................................................................................................28
2
3.1
Concept of ORP..................................................................................................................... 28
3.2
The ORP Scale....................................................................................................................... 28
3.3
Measuring the ORP Value........................................................................................................ 29
3.4
Composition of the Measuring Electrode................................................................................... 31
3.5
Composition of the Reference Electrode..................................................................................... 32
3.6
The Measuring Circuit............................................................................................................. 32
3.7
Standard ORP vs pH Compensated ORP................................................................................... 32
3.8
Standard Maintenance and Calibration..................................................................................... 33
TI12B00A20-01E First Edition March 2014
Back to the pHuture
4 Product & Features........................................................................................................36
4.1
Electronics/ Transmitters/ Converters......................................................................................... 36
4.2
Sensors/ Electrodes................................................................................................................ 39
5 Lifespan & storage of electrodes.................................................................................. 46
5.1
Recommended storage guidelines pH electrodes........................................................................ 46
5.2
Recommended storage guideline.............................................................................................. 46
6 Troubleshooting and diagnostic.....................................................................................48
6.1
Errors resulting from cracked membranes......................................................................................... 48
6.2
Fouling of the pH sensor.......................................................................................................... 48
6.3
Prevention of the reference electrode fouling.............................................................................. 49
6.4
Poisoning of the reference electrode......................................................................................... 49
6.5
Errors caused by damped or electrode cable and connection...................................................... 50
6.6
Errors caused by shorting the electrode cables........................................................................... 50
6.7
Interferences by stray voltages in the liquid................................................................................ 50
6.8
Errors caused by poor installation............................................................................................. 50
7 Frequently Asked Q&A.................................................................................................. 51
8 Appendix 1: Chemical Compatibility............................................................................. 60
9 Appendix 2: Definitions................................................................................................ 64
10 Appendix 3: Liquid Application Data sheet................................................................... 66
TI12B00A20-01E First Edition March 2014
3
1. INTRODUCTION
w w w.yo ko g a wa.c o m
1. INTRODUCTION
Measuring pH/ORP is very common, but taking true measurements and correct interpretation of the results is not self-evident.
Certain effects can potentially cause problems if not taken into consideration.
The purpose of this book is to provide a comprehensive understanding of pH/ORP measurement and how to achieve reliable
results. Basic information on the principles of measuring pH/ORP, the construction of the sensing elements and their basic use
in process applications are provided.
A part of achieving accurate and reliable pH/ORP measurements requires sufficient and correct maintenance and storage
conditions. Prevention of common errors during maintenance and storage, as well as consistent detection of loop failures is
important. This book describes how these can be avoided and how failures can be detected.
This book is accompanied with a frequently asked question and answer section as well as an appendix that includes helpful
information like a Chemical Compatibility Table and a Liquid-Application-Data-Sheet, which can be used to describe the user’s
application.
In brief: This book will make your job easier!
2. Basics on pH theory
2.1 Concept of pH
T
he Danish scientist Sørensen defined
the concept of pH as follows:
pH equals the inverse of the logarithm
to the base 10 of the hydrogen ion
concentration, as shown by the formula:
pH = -10log [H+ ] = paH1)
To facilitate the accurate measurement
of pH, and its presentation as a scale,
a range of “standard liquids” or “buffer
solutions” are used.
(1)
Later Sorensen found this definition to be
incorrect, since more concentrated solutions
appeared to give deviations between
calculated and measured values.
The definition therefore had to be modified to:
pH equals the inverse of the logarithm to
the base 10 of the hydrogen ion activity2)
as shown by the formula:
pH = -10log aH+ = pH3)
The activity of other ions present in
the solution
n
The temperature of the solution
n
The character of the solution.
n
(2)
The activity of the hydrogen ions is not
always linear with the concentration, since
this activity is not only affected
by the concentration of ions, but also by
other factors, such as:
These liquids, whose constituents are
accurately defined, have known stable
values.
Although in the preceding text the
relationship to hydrogen ions has been
made, research has shown, that the
activity of hydroxonium ions (H30+ ) is
more relevant. In aqueous solutions free
H+ ions do not occur, but are always in
combination with water molecules.
H+ + H2O ↔ H3O+
Consequently, a more correct definition for pH is:
pH = -10log aH3O+
(3)
For clarity, the notation H+ will be used in
the book as the hydroxonium ion.
Note 1. T he notation -10log .... can also be
written p ....
Note 2. See appendix 2: Definitions.
Note 3. See Chapter 2.8: Buffer solutions.
Table of Contents
4
TI12B00A20-01E First Edition March 2014
Søren Peder Lauritz Sørensen (1868-1939)
Born in Havrebjerg, Denmark, Sørensen was
a Danish chemist, famous for the introduction
of the concept of pH, a scale for measuring
acidity and basicity. From 1901 to 1938
he was head of the prestigious Carlsberg
Laboratory, Copenhagen.While working at the
Carlsberg Laboratory he studied the effect of
ion concentration on proteins,and because the
concentration of hydrogen ions was particularly
important, he introduced the pH-scale as a
simple way of expressing it in 1909.
Back to the pHuture
2.2
THe pH sCALe
Your starting point for the pH scale is pure
water which is said to be neutral. Water
dissociates1) into:
+
H2O ↔ H + OH –
(4)
Pure
or:
wate
r
2. BAsICs On pH
THeORy
Water has an equilibrium constant 2)3):
[H +] . [OH –]
Kw =
(5)
= 10 –14
[H2O]
+
-log Kw = pKw = -log [H ] + -log [OH -]
= 14 log 10
(6)
Pure water divides to give equal numbers
of H+ and OH– ions and consequently, the
concentrations of ions are 10 –7 so that:
pH = pOH = 7
Fig. 2.2a. pH value of pure water against temperature.
–
The pH value of pure water is 7.
This statement is incomplete, since the
equilibrium constant depends on the
temperature.
The definition should be: The pH value of
pure water is 7 @ 25°C.
Fig. 2.2a. and the table show the pH
variation of pure water with temperature.
If the concentration of H+ ions in a solution
is increased (e.g. to 10 –4), then the
solution has an acid character. In this case
the pH value is lower than 7.
Some examples of common solutions with
an acid character are:
H2S04 ↔ S042 + 2H
Sulphuric acid
–
+
HCI ↔ CI + H
Hydrochloric acid
+
T(ºC) pKw
pH
–
If the concentration of OH ions in a
solution is increased (e.g. to 10 –10)
then the solution is said to have a base
character.
In this case the pH value of the solution is
a number greater than 7.
Some more examples are:
NaOH ↔ Na+ + OH Caustic soda
–
NH3 + H2O↔ NH4+ + OH
Ammonia
aqueous ammonia
Note 1. See APPenDIx 2: Definitions
Note 2. The equilibrium constant is the ratio
between the rate of decomposition
and the rate of composition.
Note 3. The concentration H20 is supposed
to be 1.
pH Table
Some examples of the difference in pH
value of various liquids, foods and fruit are
shown in fig. 2.2b.
These can be compared with the pH
values of common chemical compounds
dissolved in water.
ALKALINITY
Fig. 2.2b.
Table of Contents
TI12B00A20-01E
First Edition
March 2014
5
w w w.yo ko g a wa.c o m
2.3 Measuring the pH Scale
The pH value can be measured by different
methods, e.g.:
2. Basics on pH
theory
A. Colorimetric pH measurement
B. Potentiometric pH measurement
2.3.1 Colorometric pH measurement
The principle of colorimetric determination
of the pH value is based on the pH
dependance of colour change.
Some examples are:
Some natural indicators are:
Litmus paper
When immersed in an acid medium the
paper shows red, it changes to blue in a
base medium. “pH paper” consists of paper impregnated with a suitable dye. After
immersion in the liquid to be measured the
colour of the wet paper can be compared
with a colour disc which shows the relevant
pH value for the varying shades of colour.
Red cabbage
Red cabbage is red in an acid medium
and blue/violet in a natural medium. In an
strongly basic medium the colour changes
to green.
Mushrooms will whiten considerably by
treating with vinegar (an acid). In a base
medium the mushrooms will turn brown.
2.3.2 Potentiometric pH
measurement
2.3.3 The semiconductor sensor
method (ISEFT)
The most often used pH sensing element
is a pH sensitive glass sensor. Other pH
sensors are used if a glass sensor is not
acceptable (e.g. antimon sensor, ISFET).
Accurate potentiometric pH will be
discussed in more depth in later chapters.
ISEFT is a, non-glass, ion-sensitive
semiconductor device (or transistor) used to
measure the changes in ion concentrations
within a solution. The current that passes
through the transistor will change in
response to the ion concentration change.
Table of Contents
6
TI12B00A20-01E First Edition March 2014
Back to the pHuture
2.4 PRINCIPLE OF POTENTIOMETRIC pH MEASUREMENT
The principle of potentiometric pH
measurement can be explained by Nernst’s
law.
E = Eo + RT. In [Mn+]
nF
Since the behavior of the gas Hydrogen
has a certain degree of conformity with a
metal (both have a positive ion formation),
Nernst’s law can also be applied to a
“hydrogen electrode”1) immersed into a
solution containing hydrogen ions.
The formula can be re-written as follows:
E = Eo 2) + RT. In [H+] (volt)
nF
(7)
R = Gas constant (R=8.314J/mol.K)
or:
E = RT . In [H+] (volt)
F
F = Farady number (F = 96493 C/
mol.)
n = Valency of the metal
[Mn+] = Metal ion concentration
T = Absolute temperature in Kelvin
2. Basics on pH
theory
Nernst found that a potential difference
occurs between a metal object and
a solution containing ions of the same
metal when the object is immersed in the
solution.
The potential difference E, caused by the
exchange of metal ions between metal
and liquid, was defined by Nernst as
follows:
With the constants:
E = Eo + 0,059 Ln [H+] (volt)
Eo = “Normal potential”
Walther Hermann Nernst (1864-1941)
Born in Briesen, West Prussia, in 1864. He
spent his early school years (Gymnasium) at
Graudentz, and subsequently went to the
Universities of Zurich, Berlin and Graz (Ludwig
Boltzmann and Albert von Ettinghausen),
studying physics and mathematics.
The “normal potential” is the potential
difference arising between metal and
solution when this solution contains 1 mol
Mn+/litre.
2.5 Hydrogen electrode, the basic principle
Around 1906 Max Cremer found that some
types of glass gave a potential difference
of which the magnitude depends on the
acid value of the liquid in which the glass
was immersed.
Later, Fritz Haber and Zygmunt
Klemensiewicz proved that this potential
difference, within a fixed pH range,
followed Nernst’s law in the same way as
with the so called “hydrogen electrode”.
Note 1. A
“hydrogen electrode” can be made
by coating a layer of platinum-black
on a platinum electrode and passing
a flow of hydrogen gas over it. The
presence of platinum-black results in
the hydrogen gas being adsorbed on
the electrode resulting in a so-called
“hydrogen electrode”
(see: fig. 2.5a).
Note 2. By definition, the normal potential
E of the metal “hydrogen” in a 1
normal H+ solution is 0 volt at all
temperatures.
Glass can be considered as an
“undercooled” electrolyte consisting of
an irregular structure (Si02) and a number
of other components which move in the
interspaces. These components commonly
consists of Na+ ,Ca2+ or Li+ -ions and give
an electro-balance of the glass membrane
(see figure 2.5a).
Fig. 2.5a The “hydrogen electrode”
Table of Contents
TI12B00A20-01E First Edition March 2014
7
w w w.yo ko g a wa.c o m
La+++ 115x10–10 m
Ba++ 1,35x10–10 m
2. Basics on pH
theory
Ca++ 0,99x10–10 m
Cs+ 1,69x10–10 m
Li+ 0,6x10–10 m
O– – 1,4x10–10 m
Si++++ 0,4x10–10 m
Fig. 2.5b. Texture of pH glass.
When immersed in aqueous solutions,
all types of pH glass have the particular
property to exchange the metal-ions of
+
the glass texture against the H ions in the
solution.
+
Since the H are bounded to a H2O
molecule and not free, the texture of the
siliceous acid will be defound during the
exchange by the bigger H3+ -ion.
As a result of this reaction a so-called “gellayer” will be developed on the surface
of the glass membrane. This gel-layer is
the equivalent of the metal in the Nernst’s
theory and is therefore essential for the
functioning of the glass sensor.
After one or two days the condition
reaches equilibrium and the resulting gellayer has a thickness between 10 and 40
nanometers.
This depends on several factors such
as, the composition of the glass and
the temperature in which the sensor is
immersed.
The movement of the ions will affect the
neutrality of the gel-layer. As a result, a
voltage will be developed preventing the
+
further transport of H ions.
The voltage development across the glass
membrane is generally explained by
means of the phase limit potential theory1).
The value of the voltage depends on the
concentration of the hydrogen ions in the
solution. Since this voltage cannot be
measured directly it will be necessary to
add a pH independent reference potential
in the measuring circuit. This addition
allows measurement of the potential
differences across the glass membrane.
After reaching the equilibrium the
hydrogen concentration (=activity) outside
the glass and inside the gel-layer are
equalized and consequently no transport
+
of H ions occurs. The voltage across the
glass membrane is 0 volt.
If the concentration of hydrogen ions in the
two phases differs from the concentration
in the solution, a transport of hydrogen
ions takes place.
Table of Contents
8
TI12B00A20-01E First Edition March 2014
Note 1: For clarity, other theoretical explanations
like the theories of the adsorption
potential, membrane potential and
statistic mechanic will not be explained.
Back to the pHuture
2.5.1 COMPOsITIOn OF THe gLAss eLeCTRODe
Gold
O connector
Contact
spring
Contact
spring
that is “welded” to the glass tube. The
bulb is filled with a “buffer” solution. A
reference pin is fitted and protrudes into
this liquid. The complete reference system
is completely separated from the other
parts of the electrode and is connected to
the plug of the electrode via a platinum
wire welded in glass.
Consequently, it is impossible for the buffer
liquid to penetrate the other parts of the
electrode.
As the glass membrane has a highimpedance resistance, an integral metal
screen which also carries a printed code
denoting applications, is fitted to prevent
pick-up of electrical interference.
Reference pin
Bending glass
Buffer liquid
pH sensitive
membrane
Alternatives to the standard bulb version
of the “pH sensitive glass membranes” are
available.
Yokogawa has developed electrodes with
the pH sensitive membrane designed as
follows.
- Ball Shape (Shockproof)
This is a universal electrode suitable for
most pH applications.
- Dome shape
The mechanically very strong pH
membrane (thickness approx. 1 mm) is
extremely suitable for measurements in
aggresive media.
- Flat shape
This design is used in combined sensors
for application in which solids are a
considerable component.
2. BAsICs On pH
THeORy
Metal
screen
normally the glass electrode has a bulb
shaped membrane of specific “pH glass”
Note: For a better understanding, the
construction of the glass electrode and
reference electrode are being shown as
the single electrodes before describing the
complete measuring circuit. Today these
single electrodes can be combined in
one pH sensor for the most applications,
which be shown later.
Fig. 2.5.1. The glass electrode
2.5.2 COMPOsITIOn OF THe ReFeRenCe eLeCTRODe
Gold
O connector
Metal-glass
junction
Reference pin
Cotton wool
Glass tube
Electrolyte
nernst found that the combination of
a metal and its insoluble salt in a salt
solution produces a constant mV potential.
A constant flow of electrolyte from the
electrode prevents poisoning of the
electrolyte around the reference pin.
When such a combination (known as a
reference system) is immersed directly in a
process liquid, variations may occur as a
result of other ions which may be present
in the liquid.
The reference system in the reference
electrode is joined to the gold cable
connector with a platinum pin fused in the
glass.
Furthermore, the reference system may be
poisoned by the penetration of “unwanted
ions” in the salt solution. To overcome this
problem an eletrolyte and diaphragm is
used to connect the metal/metal salt with
the process liquid (see fig.2.5.2).
By means of the reference electrode, it is
possible to measure the pH dependent
potential of the glass electrode very
accurately.
Diaphragm
Fig. 2.5.2 The reference electrode.
Note: This potential is temperature dependant.
Table of Contents
TI12B00A20-01E
First Edition
March 2014
9
w w w.yo ko g a wa.c o m
2.5.3 THE MEASURING CIRCUIT WITH A GLASS AND A REFERENCE ELECTRODE
A
E8
E7
E6
The diagrams show a pH measuring circuit
using a Yokogawa pH analyser. It consists of
A-B
B
the pH glass electrode, reference electrode,
solution ground and pH analyser built as
dual amplifier system. The diagrams show
the potentials which effect the final potential
difference (Et) between the glass electrode
and the reference electrode.
2. Basics on pH
theory
Et
E3
E4
R elecrolyte
R elecrolyte
E2
E5
R membrane
R diaphragm
E5 E4
E1 E2 E3
Fig. 2.5.3a and 2.5.3b Measuring circuit.
E1
R
liquid
The following potentials are the most significant:
E1 = potential difference between the pH sensitive glass membrane and the liquid to be measured.
E2 = potential difference between the electrolyte in the glass electrode and the inner face of the glass membrane.
E3 = potential difference between the electrode pin and the electrolyte in the glass electrode.
E4 = potential difference between the electrolyte and the electrode pin in the reference electrode.
E5 = potential difference that occurs at the interface of two liquids with different concentrations, namely the electrolyte and the process liquid.
E6 = potential difference between pH element and solution ground at Input B of dual amplifier
E7 = potential difference between reference element and solution ground at Input A dual amplifier
The total sum (Et) of these potential
differences is measured by the pH- Analyser:
E t = E1 + E2 + E3 + E4 + E5
(8)
Potential (E6) is the potential of pH electrode
against solution ground:
E6= E1-E2-E3
Potential (E7) is the potential of Reference
electrode against solution ground:
E7= E4 + E5
As we are only interested in the potential
difference between the glass membrane and
the process liquid to be measured (E1) the
remaining potentials must be compensated
for so that they do not affect the true
pH measurement. Re-examination of the
potentials generated shows that:
If the reference systems in the glass and the
reference electrode are identical and they
are at the same temperature1), then the
potentials (E3 and E4) generated by each are
equal but opposite:
E(3 )=E4 ==> E(3)-E4=0
The potentials E3 and E4 are defined as follows:
in which
Lmz = solubility product of sparingly soluble salt
Cz- = concentration of the salt solution
The equation (8) will then be simplified to:
Et= E1+ E2+ E5
(9)
With correct selection of the electrolyte used
in the reference electrode and a good flow
through the liquid junction, the potential
difference E5 can be neglected, so that
Et = E1 + E2
o
Or
E1 = E + RT. In [H+]outer
F
(10)
(11)
o
E1 = E - 0.05916 • pHouter
in the same way E2 can be defined as:
o
E1 = E - 0.05916 • pHinner
Since E1 and E2 in the pH measuring loop
are of opposite polarities the equation
becomes:
E t = E2 - E1
Et = 0.05916 (pHinner- pHouter)
(12)
the glass membrane and the process liquid.
The ideal conditions described above
cannot always be completely realised in
practice. A small potential difference may
exist when the glass and the reference
electrode are both immersed in a liquid
of similar properties and pH value to the
electrolyte.
Et = 0.05916 (pHinner- pHouter) + Easy
(13)
This potential difference is called the
asymmetric potential of the measuring system.
The asymmetric potential Easy may be caused
by:
- The liquid diffusion potential (E5 ≠ 0).
The potential difference is the result of
concentration differences across the flow
diaphragm and is called the diffusion
potential difference.
- The inner and the outer faces of the pH
sensitive membrane vary because of
differences in glass texture which occur
during the glass blowing.
The potential RT . pHinner is kept constant
F
by filling the glass electrode with an
electrolyte with good buffer properties and
consequently, the measured E1 now only
depends on the potential difference between
Table of Contents
10 TI12B00A20-01E First Edition March 2014
E1 + E2 ≠ 0
Note: 1. S ee chapter 2.6: “The effect of temperature”.
Note: 2. Errors resulting from Easy are
compensated for during calibration using buffer
solutions.
Back to the pHuture
2.5.4 MORE INFORMATION ABOUT THE pH GLASS ELECTRODE
The correct selection of a glass electrode for
a particular application can only be made if
details of the components of the measuring
loop and their significant properties are
known. The following points will be
considered in detail:
135
n
n
n
n
n
n
s election of the glass membrane
sensitivity of the glass electrode (mV/pH)
alkaline error
acid error
chemical resistance of the glass membrane
electrical resistance of the glass membrane
- High temperature
- Alkaline
100
Temperature (ºC)
70
G-Glass
- General purpose
0
0
pH
10
14
The glass membrane
reference is just as
important part of
the complete pH
measuring loop. pH
sensitive glass has the
particular property
that alkali metal ions
present in the texture
of the glass are
+
exchanged with H
ions of the liquid.
n ”G” glass
This is used for the membranes of electrodes in
processes where the nominal pH value varies
around pH 7. Since this type of glass has a
wide application range it is also termed “general
purpose” glass.
n “L” glass
The application of “L” glass is for measurements in
alkaline media with high process temperatures.
2. Basics on pH
theory
2.5.4.1 Selection
of the glass
membrane
L-Glass
To facilitate this process, it is necessary
for the pH sensitive glass membrane to be
“conditioned” by allowing it to absorb a film
of water or gel-film.
Conditioning is achieved by soaking the
electrode in water for a minimum of 24
hours.
The selection of the correct type of glass
electrode depends on both the type and
thickness of the glass membrane. Two types
of glass are available, as described here:
Fig. 2.5.4. Types of pH sensitive glass and their application ranges.
Note 1: The diagram shown in fig.2.5.4 is intended to assist with selection of the
most suitable type of glass electrode in conjunction with the application
range shown for each type. The range of any particular glass type also
depends on the membrane thickness. Three different thicknesses are
available.
2.5.4.2 Sensitivity of the glass electrode
The most important requirement in any
electrode system for pH measurement is that
the actual mV/pH ratio generated is as close
as possible to the theoretical value.
The potential generated by a glass electrode
is given by equation:
RT
E = –––– . 2,303 (pHinner – pHouter) mV
nF
long period the sensitivity will not reach an
optimal value until the electrode has been
conditioned for a number of hours.
When the decrease in sensitivity is caused
by ageing of the glass membrane the
electrode can be re-activated by etching the
surface of the glass membrane.
This should be done by immersing the electrode
for 10 seconds in a solution of vinegar (1 mol.)
and potassium fluoride (1 mol.). Ratio 1:1
At a temperature of 25°C the equation
becomes:
Note: T he electrode must be cleaned carefully
before and after activation. (For cleaning
and re-activation see: “Direction for use”
enclosed with each electrode). Frequent reactivation decreases the life of the electrode.
273+25
E = 59,16.––––– . (pHinner – pHouter) mV
298
In the equation pHouter is the pH value of
the liquid at the outer face of the glass
membrane and pHinner is the value of
the electrolyte at the inner face of the
membrane.
2.5.4.3 Alkaline of the glass electrode
In addition to a correct mV/pH ratio it is
essential that the generation of potential
difference is only influenced by the activity of
the H+ ions and not by the presence of the
other monovalent cations such as Li+, Na+,
etc.
A low concentration (or activity) of H+ions
(typically a pH value of 13) and a high
activity of alkaline ions may cause measuring
errors of between 0.5 to 1.0 pH.
In practice, the alkaline error is often caused
by sodium and consequently the term
“sodium error” is also used to describe this
effect.
Sodium Error
0
-50
-1 00
-1 50
U [mV]
The mV/pH ratio is called the sensitivity or
slope of the electrode.
The quality of the glass membrane is the
most important factor in achieving correct
electrode sensitivity.
Reduction in sensitivity of glass electrodes
may be compensated for by adjustment of
the mV/pH ratio or slope, at the analyser.
The decrease in sensitivity is usually caused
by fouling of the glass membrane. It is of
the greatest importance that the electrode
is properly cleaned before “buffering” and
adjustment for sensitivity is made.
If the electrode has been stored dry for a
Note 2: G
lass electrodes manufactured by Yokogawa are “preconditioned”
and may be used immediately without soaking. To form and
maintain the gel-film, the sensitive glass bulb is protected with a
rubber containing a small quantity of water which forms a wet
pocket for the membrane.
-2 00
U [mV]
-2 50
-3 00
-3 50
-4 00
7
9
11
13
pH
The alkaline error can be
considerable reduced by
making certain additions to
the pH sensitive glass which
improves the selectivity
of the electrode. In Fig
2.5.4.3 the alkaline error of
the different types of glass
at varying pH values are
shown.
As well as affecting
the selectivity, the
aforementioned additions
also influence other
properties, such as,
chemical resistance and
glass resistance.
Fig. 2.5.4.3 Alkaline and acid errors of the glass electrode.
Table of Contents
TI12B00A20-01E First Edition March 2014
11
w w w.yo ko g a wa.c o m
2.5.4.4 Chemical resistance of the
glass membrane
2. Basics on pH
theory
The chemical resistance of the glass
membrane is greatly influenced by the
process conditions. High temperatures and
high salt concentrations or applications in
strong alkaline liquids generally shorten the
electrode life.
Additives can be included during the
manufacture of the glass that make it
more resistant to attack and consequently
electrodes can be produced that are suitable
for measurements in either strong acid or
strong basic liquids. In aggressive solutions
a heavy duty electrode with a thick, dome
shaped, glass membrane is preferable.
2.5.4.5 Electrical resistance of the
glass membrane
Since glass is a good insulator,
potentiometric measurements cannot be
obtained with normal glass, and constituents
must be added which will reduce the
membrane resistance below 1000 MW to
minimise the effect of electrical disturbances
on the measurement.
The composition of the glass, its thickness,
the surface of the glass membrane and the
temperature, all affect the value of the glass
resistance.
Typical resistances of glass electrodes with
shock-proof bulb membranes at 25°C, are
as follows:
Type of glass
Membrane resistance
G-glass
50-100 MW
L-glass
300-500 MW
The thickness of the glass bulb of G glass
affects the electrical resistance, as follows:
Bulb shaped (shock-proof): 50-100 MW
Dome shaped (heavy duty): 120-200 MW
2.5.4.6 The response time of the
glass electrode
The response time of a glass electrode
indicates the ability of an electrode to follow
accurately any changes in the pH value.
The response time is normally defined as
the time taken to reach 63% of the value of
a step change in input. Since, in practice,
the response time depends on a lot of
factors e.g.: the reference electrode used,
the conductivity of a liquid, the temperature,
the position of the electrode in the process,
the process flow, the flow speed, etc. the
response time quoted for a particular type is
only an approximation.
Example: Glass electrode, type SM21-AG4
(shock-proof membrane).
pH change 63% of the end scale
value is reached after:
1.68 to 7
5 seconds
7 to 1.68
5 seconds
2.5.5 More information about the Reference System
2.5.5.1 General
In earlier chapters the various requirements
for glass electrodes to give accurate pH
measurements are described in detail.
The accuracy of the measurement also
depends on the properties of the reference
electrode used. It is important therefore, to
describe the different properties of reference
electrodes so that a correct selection can be
made.
The application range and specification are
shown in table 2.5.5.2.
Metal
Glass Junction
Metal
Glass Junction
A good reference electrode satisfies the
following requirements:
n
the output voltage is determined by
Nernst’s law
n
the output voltage is stable.
In the description below the different types of
reference systems, the flow diaphragms, and
the electrolytes used in reference electrodes,
are all discussed.
2.5.5.2 Reference system
Generally, the reference system used in
reference electrodes, are:
Silver/Silver chloride-Potassium chloride:
(Ag/ AgCI-KCI)
This reference system consists of a silver wire
electrolytically coated with silver chloride.
This metal - metal salt combination is dipped
Mixture
of Ag + AgCI
Cotton Wool
Silver coated with
Silver chloride
High temperature
Normal
in a potassium chloride solution (KCI).
A second type of construction for this system
consists of a silver wire dipped in a paste of
silver chloride, silver and potassium chloride.
Type of reference system
This paste is sealed into a tube by means of
a plug wadding soaked in KCI.
This reference assembly is similarly dipped in
a KCI solution.
Silver chloride wire (AgCI) in 1 molal KCI
Output voltage with regards Application range/remarks
to H2 electrode at 25°C
+223 mV ±5 mV
upto 100°C
Silver chloride paste (AgCI) in 1 molal KCI
+230 mV ±5 mV
upto 120°C
Silver-silver chloride in saturated KCI
+198 mV ±5 mV
upto 120°C
Table 2.5.5.2. Application area and specification of various references systems.
Table of Contents
12 TI12B00A20-01E First Edition March 2014
Back to the pHuture
2.5.5.3 Junctions of the reference
The selection of the correct type of junction
of a reference electrode depends on
the process conditions under which the
electrode has to function.
1
Ceramic
2
Ceramic
3
PTFE
4
Sleeve
The following junction types are available:
(see figure 2.5.6).
2. Basics on pH
theory
1.Ceramic junction.
2.Ceramic junction.
3.P.T.F.E. junction.
4.Glass sleeve capillary element.
The purpose of the junction is to maintain
contact between the reference system in the
electrode and the process liquid.
When selecting the correct junction,
consideration has to be given to ensure
that the process liquid does not penetrate
into the electrode causing poisoning and a
consequential unstable liquid junction potential.
With the first two types of junction, listed
above, the KCI solution flows slowly into the
process. The flow rate is dependent on the
over-pressure in the electrode and on the
process temperature.
The electrolyte flow rate increases with
increasing temperature.
For use in very dirty liquids a glass sleeve
capillary element is preferred because of its
larger flow surface. The sleeve can be easily
cleaned by first moving the ground ring
upwards and then wiping the ground faces.
Non-flowing reference electrodes with a
Fig. 2.5.6
porous P.T.F.E. junction can also be used
in many dirty liquid applications. The dirt
resistant properties of P.T.F.E. will prevent
complete fouling of the diaphragm.
2.5.5.4 Electrolytes in the reference
electrode
The most common electrolyte used in
reference electrodes are:
n 1moal KCL solution (with or without gel)
n 3.3 molal KCI solution
n
saturated KCI solution.
The electrolyte in the reference electrode
must satisfy the following requirements:
n
chemically inert and neutral
n
no reaction with the process liquid
n
having a constant activity of ions
n
equitransferent i.e. the ions of the
electrolyte must pass the diaphragm at
equal speed
n
having a low electrical resistance
Table of Contents
TI12B00A20-01E First Edition March 2014
13
w w w.yo ko g a wa.c o m
2.5.5.5 Pressure compensated
reference electrode
2. Basics on pH
theory
Reference
system
Bellows
Electrolyte
Process pressure
Flow
diaphragm
In processes with pressure variations, the
composition of the electrolyte may change
as a result of process liquid penetration
into the electrode.
Any change in composition of the
electrolyte may cause a measuring error or
even poisoning of the reference system of
the electrode.
The pressure inside the bellows equals the
pressure outside and only the elasticity of
the bellows itself causes the over-pressure
which results in a flow of electrolyte.
When the bellows are fully “expanded”
the electrolyte is exhausted and refilling is
required. The bellows must be compressed
before refilling.
To alleviate this problem, an electrode
with an integral pressure compensation
system (SR20-AC32) may be the solution.
See figure 2.5.5.5.
Integral pressure compensation systems
operate in a way where the electrolyte
vessel of the electrode contains bellows
which are compressed in the working
position. One side of the bellows is
connected to the pressure via the ceramic
junction and at the other side via the inner
tube.
Note: T he pressure compensated reference
electrode can also be used in processes
with pressures below atmosphere.
Fig. 2.5.5.5 Pressure compensated reference
electrode (SR20-AC32)
Gold
O connector
Metal-glass junction
Reference system
Cotton wool
Glass reservoir for
reference system
Electrolyte
Double junction
diaphragm (salt bridge)
Double junction
Electrolyte
Flow
diaphragm
Fig. 2.5.5.6 Reference electrode
with double junction (SR20-AP24)
2.5.5.6 Reference electrode with
built-in salt bridge by using double
junction
In chapter 2.5.5.4 it is explained that the
electrolyte in the reference electrode may
not be changed by penetration of the
process liquid.
Example: Mercury (Hg22+), Copper (Cu+),
Lead (Pb2+) and Silver (Ag+) ions in the
process liquid will give a reaction to the
KCI solution from the referenoe electrode.
The black is a deposit of silver sulphide in
or directly after the flow diaphragm.
The results of such deposits can be:
n
long response of the pH measuring
circuit
n
non-reproducible diffusion voltages and
consequently drift in the indication.
n
calibration is hardly possible (the formed
diffusion voltage can be pH dependent).
n
increased resistance of the diaphragm
(slower measurement).
To solve this problem the KCI solution and
the process liquid must be separated using
a double junction electrolyte; resulting
in a reference electrode with a built-in
saltbridge.
Processes containing cyanides, bromides,
iodides or sulphides are a second
example of selecting the KCI solution
critically.
Mostly, a black diaphgram indicates that
the reference electrodes is used without a
double junction.
Table of Contents
14 TI12B00A20-01E First Edition March 2014
Note: M
ost biological process liquids contain
sulphuric compounds.
Back to the pHuture
2.5.6 COnsTRUCTIOn OF THe TeMPeRATURe eLeCTRODe
pH measurements are temperature
dependent from two different effects:
a. by the variations with temperature on
the contact potentials in the glass and
the reference system.
b. by temperature variations of the liquid
being measured.
Gold
O connector
2. BAsICs On pH
THeORy
Screen plate
As an alternative to automatic temperature
compensation it is possible to manually
compensate for temperature variations.
Therefore it is necessary to include
a temperature compensator in the
system, whose purpose is to provide
automatic compensation for the effects of
temperature variations on the measuring
system.
Tube
Pt100
resistor
Thermal
conducting grease
These compensators are made in the same
shape as the other electrodes to enable
them to be mounted in the same fittings.
The temperature compensator consists of
a platinum resistance element (e.g. PT100,
PT1000), mounted in a glass tube. The
tube is completely filled with white silicon
grease, the thermal conducting properties
of which ensure fast temperature response.
Fig. 2.5.6. The resistance thermometer
2.5.7 THe COMBIneD pH sensORs
In today’s business we see a tendency
to use combined sensors instead of
separate electrodes. In modern combined
electrodes the glass-, reference-,
temperature electrode and solution ground
are built into one unit.
Wide body sensor
Glass sensor
The reference systems of both the glass
and the reference electrode consist of an
Ag/ AgCI with same KCL solution.
The operating principles are identical to
those used for the individual electrodes.
The advantage of combined sensors is
easy maintenance.
2
2
1
9
3
1
4
9
4
6
10
7 8
6
5
1
2
3
4
3
10
Reference electrode
Reference element
Electrolyt
Juction
5
6
7
8
7 8
5
pH glass electrode
Internal reference element
9 Temperature element
Glass membrane
Internal buffer solution 10 Solution ground
There has been a progression in the
design of combined sensors from the
ability to have just a pH and reference
in a 12 mm design, to also incorporate
the temperature element and the solution
ground. Yokogawa has managed to fit
it all electrodes into the 12 mm design
with the development of the SC24V and
SC25V sensors.
By incorporating the solution ground the
possibilities for performance improve,
along with predictive maintenance and
diagnostic capabilities have improved.
Fig. 2.5.7. The combined pH sensors
Table of Contents
Remark
TI12B00A20-01E
4 Wide body sensor should be simplification
of FU20
Glass sensor should be the simplification of SC25
First Edition
March 2014
15
w w w.yo ko g a wa.c o m
2. Basics on pH
theory
2.6 The effect of temperature
2.6.1 Temperature effect on the glass
and the reference electrode
The glass and the reference electrodes
have a number of temperature dependent
contact potentials; it is obvious then that
the voltage supplied by the measuring
system is temperature dependent.
This temperature dependancy is shown by
the factor in the
Nernst equation RT
F
The voltage supplied by the measuring
system is:
E t = Easy x 2,303 RT .(pHinner – pHouter)
F
E t = Easy x 59,16 ( T+273 ) (pHinner – pHouter)
298
mV
–500 –400 –300 –200 –100
60ºC
20ºC
0
1
2
3
4
5
6
7
S
8
9
10
11
12
13
14
T is the temperature in °C
If the glass and the reference electrodes
are immersed in liquids of equal
temperatures, the potential variations of
similar reference systems will be equal and
opposite.
E3 = -E4
Consequently, the system will be
unaffected by temperature variations.
The temperature effect on the contact
potential of the junction on the reference
electrode is kept to a minimum by correct
selection of the junction and electrolyte.
The temperature effects obtained by immersing the electrodes in different standard solutions and then by varying the temperature of
these standard solutions, are shown in the
graphs of fig. 2.6.
0ºC
mV
80ºC
pH
Fig. 2.6 Temperature effect on the mV/pH ratio.
This graph shows that:
a.the mV/pH ratio increases as the
pHinner is standardised at pH 7.
+100 +200 +300 +400 +500
temperature of the measuring system
increases.
At 25°C the mV/pH ratio is 59.16 and
at 20°C this ratio is 58.16 mV/pH.
At 80°C the mV voltage per pH unit is
increased to 70.08 mV.
b.the various isothermal lines intersect
In general, when a pH measurement is
made in a process at widely fluctuating
pH and temperature levels, automatic
temperature compensation is necessary.
To achieve this the electrode system is
completed with a temperature sensing
elements, packaged in a similar
construction to an electrode, that
compensates for slope variations of the
mV/pH ratio of the electrode system.
at one point S (the isothermal point of
intersection)
c.tthe intersection point is dependent on
the pH of the buffer solution used in the
glass electrode (this is usually pH 7).
It is important that the isothermal lines
intersect at only one point. So selection
of the correct buffer solution is essential
in order to obtain an accurate isothermal
point of intersection S, shown in figure
2.6.
Note 1.The isothermal point of intersection of
the standard electrodes of Yokogawa
is at pH 7. Depending on the buffer
solution used this point may, for special
applications be at another value pH 3.
2.6.2 Temperature effect on the process liquid
In the preceding chapter, the temperature
effect on the measuring system and its
correction, has been considered. In addition
there is however, a temperature effect on the
chemical balance of a process itself.
It has been previously stated that the pH
value of pure water at 0°C differs from the
pH value at 100°C.
This is caused by a change of the chemical
balance.
H2O ↔ H+ + OH at 0°C for pure water
pK = 14,94 or pH = pOH = 7,47
at 100°C for pure water
pK = 12,24 or pH = pOH = 6,12
For accurate comparison of pH values
made by different techniques (e.g. by lab.
measurement and industrial measurement), it
is necessary to state at what temperature the
measurement was made.
As the effect of temperature on any process
liquid is highly dependent on its composition,
it is not possible to accurately compensate
for this effect automatically.
Table of Contents
16 TI12B00A20-01E First Edition March 2014
Back to the pHuture
2.6.3 Temperature effect on the application range of the glass electrode
10000
AG2 (25M)
Glass Impedance
1000
AG4 (100M)
MOhm
100
AG6 (200M)
10
1
AL6 (1000M)
where:
HIGH LIMIT
0.1
The following relation can be derived
0.01 (at reference-temperature):
-5
c H53O +15� d25 . 35
c OH45� �d55
(
)(
LOW LIMIT
65
75
85
) = Kw
degC @ ref .temp
95
105 115 125
Fig. 2.6.3 where:
Temperature effect on 4471
the glass
.33 membranes. resistance.
� A .......
+ 0.017053 (t + 273.15 ) � 6.0846
t + 273.15
Where: t, ref-temp = temperature expressed in: oC
Kwt = 10
A=
2. Basics on pH
theory
Process temperature is a major factor in the
selection of the type of glass electrode to be
used for a particular application. Different
reference systems are used for high or low
process temperatures.
Furthermore, the chemical resistance of the
glass membrane is temperature dependent
and correct selection is important.
A third factor is the membrane resistance
of the glass electrode. This increases
considerably at lower process temperatures
and may increase the response time to an
unacceptable level.
A rough guide is that the glass membrane
resistance increases by a factor 2 with
every temperature fall of 10°C. Figure 2.6.3
shows the resistance of glass membranes for
various species of glass.
The compensated pH value is:
(
pH ref = � log 10 � pH � d
)
formula: 3.1.1.1
2.6.4 NEN6411 Temperature compensation
Matrix
where d: = concentration change
for d, the next equation can be solved:
Using the NEN6411 norm temperature
compensation can be calculated and is
applicable for many applications.
It’s used for pH compensation in water
applications using a glass-electrode.
The calculation is valid for all strong acids
and strong bases. The main application is
in de-mineralized water and alkalised boiler
feed water/condensate.
The uncompensated pH value is:
10 � pH +
d=
Kw@ temp
10
� pH
�
10 � pH +
Kw@ temp
10
� pH
2
� 4 . (Kw@ temp � Kw@ ref .temp )
2
formula: 3.1.2.2
Where:
Kw @temp = f (Temp)
Kw @ref -temp = f (RefTemp)
NEN 6411 pH Temperature Relationship for Strong Acids and Bases
where:
The following relation can be derived
(at reference-temperature):
( c H 3O +
� d ) . ( c OH � �d
) = Kw@ ref .temp
where:
4471.33
Kwt = 10 � A ....... A =
+ 0.017053 . (t + 273.15 ) � 6.0846
t + 273.15
Where: t, ref-temp = temperature expressed in: oC
The compensated pH value is:
(
pH ref = � log 10 � pH � d
)
formula: 3.1.1.1
where d: = concentration change
for d, the next equation can be solved:
10
d=
� pH
+
Kw@ temp
10 � pH
�
10
� pH
+
Kw@ temp
10 � pH
2
pH at Reference Temperature
10
9.5
80 ºC
9
8.5
0 ºC
8
7.5
T = Tref = 25ºC
7
6.5
6
5.5
5
� 4 . (Kw@ temp � Kw@ ref .temp )
5
2
7
8
9
pH at Process Temperature
formula: 3.1.2.2
Where:
Kw @temp = f (Temp)
Kw @ref -temp = f (RefTemp)
6
Fig. 2.6.4
Table of Contents
TI12B00A20-01E First Edition March 2014
17
w w w.yo ko g a wa.c o m
2. Basics on pH
theory
2.7 Isolation resistance
highest order. In industrial applications
the analyser should be installed as near
to the electrodes as possible. At all times
the connections between electrodes and
analyzer should be kept dry. The insulation
resistance decreases considerably when
any moisture is present.
The insulation resistance of the reference
electrode is less criticall as its resistance
with respect to the measuring liquid is
much lower. Generally, an insulation
resistance of 107W is adequate1).
The resistance between reference
electrode and liquid is usually between 1
and 10kW at 25°C, depending on the
type of junction.
At higher resistance values the sensitivity of
the measuring system will be reduced and
may cause an instability of measurement.
Buffer solutions are needed as
indispensable tool for maintaining
an accurate pH measurement. Buffer
solutions are used as references points
for calibration and adjustment of pH
measurements to compensate ageing and
deterioration.
Buffer solutions are mixtures of weak
acids and the salt of these acids with a
strong base, or mixtures of weak bases
and the salt of these bases with a strong
acid. Consequently, if the buffers are not
accurate themselves, the calibration serves
no useful purpose.
Buffers are classified in three categories.
The main difference between the different
types of buffers is the accuracy and buffer
capacity.
Primary reference buffer
These buffers are not commercial buffer
and mainly used in metrological institutes.
These buffers show the lowest uncertainty
in pH values, ±0.003.
Standard Buffer (secondary reference
buffer)
Standard buffer solutions are used as
standards for accurate measurements
especially in laboratories and production
of technical buffers. They are traceable to
the primary standards. The constituents of
these buffers are defined by international
standards like DIN19266, IEC 726 and
NIST.
The uncertainty is 0.002 and 0.004 pH
units (at 25°C), depending on the buffer
Technical buffer
They are commercial buffers and used
mainly for calibration of industrial pH
measurements.
The buffer values of technical buffers
are traceable to standard buffer. The
DIN19267 defines standards for these
solutions. The uncertainty is 0.02 a pH
units (at 25 °C), depending on the buffer
Examples of preferred buffer by Yokogawa
are shown in the table below. Buffer
solutions prepared from these substances
conform to the recommendations of
the DIN Standards Committee and
the National Institute of Standards and
Technology (NIST). The substances were
chosen for their particular suitability as
calibration standards for precision pH
meters.
Temperature dependence
for temperature variations will be done. Any
stated pH value is only meaningful if the
measuring temperature is also specified.
In view of the relatively high resistance
of the pH sensitive glass membrane, it is
necessary to use a analyser with a high
input impedance. This impedance must be
at least a factor of 1000 higher than the
membrane resistance (the resistance of the
reference electrode is much lower and can
be neglected).
Insulation and screening of all cables
and connections between the measuring
electrode and the analyser must be of the
Note 1:pH measurements in low conductivity
liquids with a analyser with two high
input impedances for both the glass and
the reference electrode require a good
insulation resistance.
2.8 Buffer solutions
The temperature dependence of the pH
of a buffer solution is generally specified
in terms of measured pH values at certain
discrete temperatures.
Many buffer tables are pre-programmed in
Yokogawa Analyzers. So if during calibration
the temperature compensator is immersed
in the buffer liquid, an automatic adjustment
Note 1: N
.B.S. National Bureau of Standards
of the U.S.A.
Note 2,3: See appendix 2: Definitions.
STANDARD BUFFER SOLUTIONS1)
Compositions
Molarity
pH
Dilution value Buffer- Temp. coeff.
at 25°C (pH1/2)2) capacity3) dpH/dT
Potassium trihydrogen dioxalate
(Tetroxalate)
KH3(C204)2• 2H2O
0.0496
1.679
+0,186
0,070 +0,0010
0.00997
9.180
+0,010
0.200 +0.0082
Potassium dihydrogen phosphate+
0.02490+ 6.865
Disodium hydrogen phosphate
+0.080
0.029
+0.052
0.016 +0.0012
Borax Na2B407 • 10H2O
Na2HP04 • 2H20 + KH2P04
Potassium hydrogen phtalate
KHC8H404
-0.0028
0.02490
0.05
Be Aware
Buffers with a pH above 7 are particularly
sensitive to atmospheric CO2. Buffer
showing any sign of turbidity must be
discarded immediately.
4.008
For accuracy it is recommended that a
buffer should not be used for more than
a month after opening. Buffers should be
stored in tightly sealed, preferably air-tight
bottles made of polyethene or borosilicate
glass. Buffers should not be returned to the
bottles once removed.
Table of Contents
18 TI12B00A20-01E First Edition March 2014
Back to the pHuture
2.9 Periodic maintenance and calibration of pH Sensors
2.9.1 Why Is Maintenance Needed?
The validiation control chart (Fig. 2.9.1)
shows that frequency of maintenance
of your measurement depends on the
required accuracy. A pH measurement
was checked daily in buffer solution
without adjustment. This chart shows
that the reading measurementis swinging
around the calibration value. To guarantee
an accuracy of 5% you have to calibrate
the measurement at minimum twice a
week. If accuracy of 10 % is accepted
you can prolong the frequency to once in
two weeks.
Periodic calibration is necessary to ensure
the highest measurement accuracy.
Calibration adjusts for the aging of the
sensors and the non-recoverable changes
to the electrodes that take place.
These effects usually happen slowly
therefore, calibration should not be
necessary more frequently than about
once a month in typical general purpose
applications. If more frequent calibration
is needed, it is usually because the
cleaning process was not effective, the
calibration was not well executed, the pH
readings are temperature dependent or
the wrong electrodes have been selected.
If a film remains on the pH sensor after
cleaning, then a measuring error can be
interpreted as a need for re-calibration.
Since, these changes are reversible with
proper cleaning, it is a key step in the
maintenance process.
Note: T he periodic maintenance advice that
follows is intentionally general in nature
because pH sensor maintenance is
highly application specific.
2. Basics on pH
theory
The selection of pH electrodes and
holders (fittings) is based on the demands
of the application where they will be
used. The desire is to achieve an accurate,
reliable measurement with a reasonable
life expectancy while minimizing the
required routine maintenance. When a
quality pH sensor system is undamaged,
clean and properly calibrated, it will
provide a measurement that is accurate
and reliable. This sounds simple enough,
but ensuring the system is clean and
calibrated will sometimes involve a
significant amount of maintenance. The
effect of dirty or faulty electrodes can be
anything from slow response to completely
erroneous measurements.
Validation control chart
Upper Control
Level (+10%)
Upper Warning
Level (+5%)
Calibration
Value
Lower Warning
Level (-5%)
Lower Control
Level (-10%)
Days
1
3
5
7
8
11 13
15
17
19
21
23
25 27
29
Fig. 2.9.1 Validation control chart
Table of Contents
TI12B00A20-01E First Edition March 2014
19
w w w.yo ko g a wa.c o m
2. Basics on pH theory
2.9.2 General Maintenance Procedure - Clean – Check – Calibrate
Clean:
The starting point for any pH maintenance
activity is to make sure the electrode
system is properly cleaned. A variety of
cleaning solutions can be used depending
on the coating effects of the process on
the electrodes. Typically a 5 to 10%
solution of HCl works well.
1) Rinse off the sensor in just tap water to
remove any heavy process coating.
2) Immerse the electrodes in the cleaning
solution for 1-2 minutes, agitating them
regularly. Use a soft brush to clean off
coating deposits without damaging the
electrode.
3) The electrodes must then be rinsed
thoroughly with tap water to avoid
contamination of the calibration solution
with the cleaning chemicals.
Check:
Often measurement inaccuracy is due to
the process coating of the electrodes and
calibration is not necessary, only cleaning.
Therefore, after rinsing off the cleaning
solution thoroughly with tap water,
immerse the electrodes in a buffer solution
and allow them to stabilize. Check the
reading! Rinse again and place the
electrodes in a second (different value)
solution and allow them to stabilize. If the
readings in both cases are within e.g. +/0.1 pH, the electrodes can be put back
on line. If not, then proceed to calibrate.
Calibrate:
A pH measurement loop requires regular
calibration of the electrodes to compensate
for their aging and deterioration. Typically
this is done about once a month. Always
use fresh buffer solutions to avoid the
possibility of introducing errors from
contaminated or aged solutions. Buffers
supplied as liquids have a limited shelf
life, especially alkaline buffers, which can
absorb CO2 from the air.
Yokogawa strongly recommends to use
buffer solution according NIST or DIN
standards in order to ensure the best
accuracy and best buffer capacity is
available. For most applications we
advise to do a two point calibration.
2A
djust the meter reading to the pH value
of the buffer solution according to the
method selected (automatic or manual).
3 Rinse the electrodes with tap water
to remove the traces of the first buffer.
Then, immerse the electrodes in the
second buffer solution (e.g. 4.01 or
9.18 depending on the measuring
range) and repeat step 1.
Note: A
lkine buffers are often inaccurate due to
absorbing CO2 during storage after first
use. So a calibration with acidic buffers
should be preferred for all calibrations to
avoid calibration errors due to aging of
buffers.
4A
djust the analyser reading to the pH
value of the buffer solution according
to the method selected (automatic or
manual).
1R
inse the electrodes thoroughly with tap
water to remove any remaining cleaning
solution. Then, immerse the electrodes
in the first buffer solution (usually the
neutral buffer) and stir the electrode
for a few seconds to ensure that the
gel layer of the pH glass membrane is
completely wetted by buffer solution and
all other fluids are removed. Then keep
the electrode still let the measurement
stabilize. For a good calibration we
advice to avoid any contact of the glass
membrane with bottom of the calibration
vessel.
A check for correct calibration have to be
done by immersing the sensor again in
the first buffer solution (after rinsing) to see
if the reading is accurate. If it is not, the
calibration should be repeated.
WARNING: During calibration the solution
ground and the temperature compensator must
be connected. The buffer solution temperature
must be within the technical specification limits
as indicated on the label.
2.9.3 Two-Point Calibration
pH 4 or pH 9). The slope is adjusted so
the line is rotated from “2” to “3”. Like this,
the zero point is adjusted to pH 7 based
Sensor voltage
mV
on the EMF of a pH sensor and the span
is adjusted with reference to the difference
from pH 7 in the pH analyser.
1
2
3
x mV
l
rea
The electromotive force (EMF) of a pH
sensor under ideal conditions is expressed
by line “3” in Figure 2.9.3. In practice, the
electrode shows the characteristic that is
expressed by line “1” because of different
properties or aging of the pH sensor. To
correct this, a zero adjustment (asymmetry
potential adjustment) and a span
adjustment (potential slope adjustment)
by the pH analyser are required. First,
re
perform
al a zero point adjustment using a
standard
solution with a pH value close
to pH 7.
0
7
pH
The line is shifted from “1” to “2” laterally
so
id it passes through the zero point.
ea perform a span adjustment using a
Next,
l
standard solution with a span pH (typically
pH
id
ea
Fig. 2.9.3 Two-Point calibration
Table of Contents
20 TI12B00A20-01E First Edition March 2014
l
Back to the pHuture
2.9.4 One-Point Calibration
Sensor voltage
mV
x mV
0
7 c a
al fte
ib r
ra
tio
n
ca be
lib for
ra e
tio
n
pH
2. Basics on pH
theory
As a simple calibration method, one-point
calibration is available. The rate of change
in generated EMF per pH (potential
slope) of a glass electrode is generally
small compared to that in the asymmetry
potential. In one-point calibration, the
adjustment of potential slope performed
in two-point calibration is omitted. There
are two ways to perform one-point
calibration. One is, as shown in Figure
2.9.4, performed by using one point of an
appropriate pH standard solution: grab
sample. The other is by using a sample
solution being measured. The pH of the
sample solution is manually determined by,
for example, a portable pH analyser, and
then the pH analyser is adjusted so that it
reads the same pH value.
Fig. 2.9.4 One-Point calibration
2.9.5 3-Ways to Calibrate
Automatic calibration
Manual calibration
Sample calibration
Yokogawa analysers offer internally
programmed buffer tables, to calculate
the buffer value at the actual temperature
during the calibration. In addition, the
stability of the reading is automatically
calculated, and when the reading has
stabilized fully automatic adjustments of
slope and asymmetry are made. This
eliminates the question of how long the
operator should allow prior to adjustment.
A menu driven prompt system conducts
the operator through the simple, foolproof
routine.
Yokogawa analyser also offers this
method, where the operator decides
the actual pH value to enter. Manual
calibration is most often used for singlepoint adjustment of the asymmetry
potential, by comparison method (grab
sample). Manual calibration can also be
used to perform a full 2-point calibration
with solutions other than the NIST buffers
that are listed in the calibration tables.
In this case, the solutions are applied
sequentially as in the AUTOCAL method,
but the user determines the adjustment of
reading and stability.
The operator activates the “SAMPLE”
calibration routine, at the same time as
taking a representative process sample.
After determining the pH of this sample
by independent methods, (in the lab for
example) the reading can be adjusted in
the analyser. While the sample is being
analysed, the analyser holds the sample
data in memory, while continuing to
control and read pH normally.
Table of Contents
TI12B00A20-01E First Edition March 2014
21
w w w.yo ko g a wa.c o m
2.9.6 Calibration Parameters
2. Basics on pH
theory
Asymmetry Potential/Zero Point
The Asymmetry Potential (AS) also referred
to as the millivolt offset, is an indication
of the condition of reference electrode of
a pH sensor. Theoretically when the pH
sensor is placed in same buffer like the
internal buffer (normally, pH7), the millivolt
output from the sensor (pH and reference)
should be zero. Causes of the millivolt
offset are:
• depletion of the Potassium Chloride
(KCl) from the reference electrolyte
•or the reference electrolyte becomes
poisoned with the process solution.
When the millivolt offset is greater than
+/- 30mV, it is advisable to replace
the reference electrode. The correction
limits of the Yokogawa pH analysers are
+/-120mV. Outside this range an error
message will appear on the display.
Often the Zero Point instead of Asymmetry
potential is used to indicate the condition
of the reference electrode. The Zero Point
shows the pH value when the sensor
voltage is zero.
Slope:
The Slope (SL) also referred to as the
efficiency of the pH sensor is an indication
of the condition of the measuring (glass)
electrode. The slope is displayed in a
percentage (%) value, with 100% SL
being ideal. When the electrode is new,
the slope should be in the upper 90%
range. As the electrode ages and loses
efficiency, the slope and response of the
electrode will start to decrease. The slope
value is updated each time a two point
500
∆pH
∆mV
mV
Asymmetry Potential
0
-200
ITP Isothermal-Point
0 mV
0ºC
10ºC
25ºC
0
7
Zero
Point
ITP Null
Fig. 2.9.6 Calibration parameters
calibration is performed and usually only
small changes in the slope value should
be noticed. One of the main reasons the
electrode may have a low slope value,
is because it was not cleaned before
calibration, making it less efficient. If a
low slope value is noticed, clean the
electrode with a 5-10% HCl solution for
a minute, rinse it thoroughly with clean
water and recalibrate. Depending on the
application and response time required, it
is advisable to replace the pH electrode
when the slope value is in the mid to low
80% range. The factory default values are
70% - 110% of theoretical and are user
programmable from the Service level.
When the calculated slope exceeds the
programmed values, an error message
will appear on the display.
pH
14
Reference Impedance:
The Reference Impedance (RZ) also
referred to as the resistance of the
reference junction, is an indication of
a precipitate (blockage) forming in the
reference junction and that the electrode
needs cleaning. This resistance is also
influenced by the conductivity of the
process solution. Typically a clean
reference junction will have a resistance
of less than 10-15kΩ , but in low
conductivity solutions, RZ values between
200 and 500 KΩ are not uncommon.
When the RZ value starts to approach 30
-35 KΩ, the electrode will start to have a
slow upward drift. When the reference
impedance exceeds 100 KΩ an error
message will appear on the display. The
RZ value is user programmable between
50 - 999 KΩ.
Table of Contents
22 TI12B00A20-01E First Edition March 2014
Back to the pHuture
2.10 DIFFeRenTIAL eLeCTRODes
changes over time by diffusion, so regular
calibration is required. Also the process
may contain ions that poison the reference
element and these ions penetrate the
element by diffusion especially at high
temperatures.
This means most of the problems are not
related to the pH measuring electrode,
but almost exclusively to the reference
electrode. Therefore the solution of the
problem must be sought in alternative
ways of generating a reference voltage
that is stable over time and is independent
on the pH value.
still incorporates a pH sensitive measuring
element to detect ph changes. However,
for generating a stable reference voltage
now a cation sensitive measuring element
is used.
The benefit of using a caption reference
electrode that it has NO junction, there is
NO path from the process to the internal
element; so NO poisoning can occur.
Also since there is NO junction, there
is NO plugging or coating problems to
worry about and there is NO electrolyte
depletion problem, because there is NO
electrolyte.
2. BAsICs On pH
THeORy
A pH sensor measures the voltage that
the pH membrane measures as function
of the pH value of the process sample.
This voltage is then compared with
the mV output of a reference cell that
is independent on the pH value of the
sensor. In a conventional pH measuring
sensor this reference electrode is like
described before a Metal/ Insoluble
metal salt/salt solution combination
inserted in an electrolyte solution (or gel)
that is separated from the process with a
junction. This assures that the electrolyte
composition does not vary with process
changes and therefore the mV output does
not change either.
However the reference electrode is still in
open electrolytic contact with the process
and the composition of the electrolyte
The best solution is to have a hermetically
sealed reference electrode, like a
differential sensor. A differential sensor
2.10.1 DIFFeRenT COnCePTs OF DIFFeRenTIAL sensORs
Different concepts of differential sensors are
known.
One type of differential sensors is using
a second pH-sensitive glass as reference
electrode (fig. 2.10.1). This glass membrane
is installed in a pH buffer solution. The buffer
solution is through a diaphragm connected
to the process. The reference potential is
created between glass membrane and buffer
solution. The advantage of the concept is
that the inner reference element will not be
poisoned by earth currents or other reasons.
But the buffer solution is still in contact with
the process and will be consumed.
Another example of a differential pH sensor
is the Yokogawa SC24V; which is a 12mm
combination pH glass electrode with a
salt sensitive glass reference electrode
membrane. In the following chapters explain
how a salt sensitive reference works and
what the benefits are.
To increase the lifetime of this buffer solution
concept, sometimes a salt bridge will be
installed between the process and buffer
solution. With this concept longer life times
are possible compared to conventional
electrodes, but we have a continuous drift
of the asymmetry potential and frequent
calibration is still necessary.
pH membrane for
measurement
chamber filled
with buffer
solution
PH18 sensor uses a pH sensitive and cation
sensitive enamel that are bonded directly to
the metal shaft with no electrolyte needed.
The Pfaudler PH18 sensor is sold from
several Yokogawa affiliates.
pH membrane as
reference
The other type of differential sensors is using
a salt sensitive membrane as a reference
element to generate the reference signal.
The reference electrode is in direct contact
with the process. The well-known Pfaudler
Fig. 2.10.1 Differential sensor with pH
sensitive sensors
Table of Contents
TI12B00A20-01E
First Edition
March 2014
23
w w w.yo ko g a wa.c o m
2.10.2 How does a salt sensitive reference work, and what are the advantages
2. Basics on pH
theory
The SC24V (Fig. 2.10.2a) is a combined
differential pH glass sensor. The sensor
uses as reference cell with a salt sensitive
glass membrane. The sensor also includes
a PT1000 temperature element and with a
Platinum solution ground pin.
Fig. 2.10.2a
sensor output of pNa sensor
100
mV vs SCE
50
0
mV
In the past Yokogawa developed this
glass as a single pNa- ion sensitive
electrode to measure the sodium ion
concentration within a process. It was
found that just as a pH measurement
sensor, the analog signal to the output of
the reference membrane follows the Nernst
Law. Fig. 2.10.2b.
-50
-100
-150
-200
-250
-300
-2
0
2
4
6
pNa
Fig. 2.10.2b
However the output voltage of the salt
sensitive reference membrane depends on
the salt concentration of the sample.
As it turns out the pNa membrane also
shows sensitivity to other cations, i.e. Li+,
K+, and Ca2+. (Fig. 2.10.2c) within a
process; so the use of this reference glass
is not limited to only sodium ions. It can be
used as salt sensitive reference in general
in the concentration is high enough and
consistent to generate a stable mV output
reading.
Sensitivity reference glass for cat-ions
Potential(mV)
0
-50
Na+
-100
K+
-150
Ca2+
-200
Li+
-250
-300
0
1
Na+
2
K+
Fig. 2.10.2c
Table of Contents
24 TI12B00A20-01E First Edition March 2014
3
Li+
4
Ca2+
-log(Kation+)
Back to the pHuture
500
pH glass and
salt reference
elements
400
300
pH glass element
mV
200
100
0
pH
-100
600 g / l
-200
60 g / l
6 g/ l
-300
0, 6 g / l
-400
salt reference element
0
2
4
6
8
10
pH
12
14
16
2. Basics on pH
theory
The graph (Fig. 2.10.2d) shows that output
signal from of the pH glass element and
of the salt reference element. The graph
shows very nice when the output signal of
salt reference is stable and independent
from changes of pH concentration and
can be used of pH measurement.
Fig. 2.10.2d
It is important to know that the salt
sensitive glass has sensitivity to H+ ions.
The usable range (Fig. 2.10.2e) indicates
the area where the influence of H+ on the
output is neglectable. Also a minimal Na+
concentration of 0.0001M is required to
guarantee a fast response.
[Na+]
pNa
NaCl saturation level
1M
0
Applicable range
0.01M
2
0.000M
4
0
2
4
6
pH
8
10
12
14
Fig. 2.10.2e: Application range for a salt sensitive reference electrode in pH measurement
The following graph (Fig. 2.10.2f) shows
the sensor output of the pH membrane
versus the reference membrane as function
of the pH value in a 1 mol brine solution
(60 g/l NaCl)
This graph shows that in this application
the sensor behaves like any normal pH
sensor with the Isopotential at pH 7, 0 mV
and a linear response over the complete
pH range down to pH 2. Below 2 pH we
see a slight non linearity. Between 0 and
1 pH the sensor sensitivity is too low to
get a good measurement.
Differential pH at 60 g/l NaCl
pH- SCE
pNa-SCE
diff
mV
450
a
400
350
b
300
250
200
150
100
50
c
0
0
1
2
Fig. 2.10.2f: Differential pH at 60 g/l NaCl
3
pH
4
5
Differential pH measurement at 20 ºC
Table
mV
of Contents
6
7
pH- SCE
pNa-SCE
diff
450
TI12B00A20-01E First
Edition March 2014
400
8
25
200
150
w w w.yo ko g a wa.c o m
100
50
c
0
0
In concentrated brine solutions (Fig.
2.10.2g) the naCl concentration is much
higher and therefore the output of the pna
electrode is different.
1
2
3
4
pH
5
6
7
Differential pH measurement at 20 ºC
8
pH- SCE
pNa-SCE
diff
mV
450
2. BAsICs On pH
THeORy
400
The sensor output is linear over almost the
complete pH range with only a slight loss
of sensitivity below 1 pH. In this scenario
the Asymmetry Potential is now -50 mV
at pH 7, instead of the tradition 0 mV. In
this situation it is good practice to the ITP
setting of the pH analyser to 5.6 pH to
reflect the application ITP. Therefore more
accurate temperature compensation is
achieved and the ASY is 0 mV.
350
300
250
200
150
100
50
0
-50
-100
0
1
2
3
4
5
6
7
8
pH
Fig. 2.10.2g: Differential pH in concentrated brine solution
There are also applications where the
salt concentration is much lower, so the
following example specifies the sensor
output at 6 g/l naCl.
This graph (Fig. 2.10.2h), shows that the
sensor is only linear down to 3 pH and
the Isopotential value is now 8 pH or the
Asymmetry Potential at pH 7 is + 60 mV.
These examples are showing that the
Asymmetry Potential or ITP are different
depending on the salt concentration of the
process.
This can be seen as a problem, because
in most applications this concentration is
not constant.
However there is no reason for worry,
because a change in salt concentration
of +/- 25% of salt only causes and offset
of the sensor output of 5 mV or an error
on the pH readings of 0,1 pH. In most
pH control applications this will not be
noticed, even when the salt concentration
varies by 25%.
Differential pH at 6 g/l NaCl
mV
400
350
300
250
200
150
100
50
0
-50
-100
0
1
2
3
4
pH
Fig. 2.10.2h: Differential pH at 6g/l NaCl
Table of Contents
26
TI12B00A20-01E
First Edition
March 2014
5
6
7
Back to the pHuture
2.10.3 Calibration
Refer to the analyser Instruction Manual for details.
Standard buffers can be used for functionally check of sensors, however if they are used
for calibration, the Asymmetry and Slope values will be incorrectly calculated.
Therefore specific buffers for sodium-based reference are available:
pH 2.00 (Ionic strength
n pH 4.00 (Ionic strength
n pH 7.00 (Ionic strength
n pH 9.00 (Ionic strength
n
–
–
–
–
1 mol NaCl)
1 mol NaCl)
1 mol NaCl)
1 mol NaCl)
2. Basics on pH
theory
Normally the pH standards that are preprogrammed in the pH analyser can be
used for calibration with the pH analyser
set to “AUTOCAL”. The SC24V however is
a differential pH sensor which needs pH
buffers that have the same ionic strength
because the sodium reference will change
as the ionic strength changes. These pH
buffers are not pre-programmed in the pH
analyser, which means the calibration has
to be done with specific buffer solutions
and the pH analyser set to “MANUAL”
calibration.
2.10.4 Buffer calibration
To calibrate the SC24V sensor, two
buffer solutions with known pH values
are required. It is recommended that one
buffer solution has a value near to pH
7.00. Depending on the process value to
be measured, the second buffer solution
should be either acidic (below pH 7.00)
or alkaline (above pH 7.00). Buffers which
are available are: pH 2.00, pH 4.00, pH
7.00 and pH 9.00.
The following is a very general 2-point
manual
calibration procedure with buffer solution:
1. Clean the sensor using a 5% solution of
HCl;
2. Rinse sensor thoroughly with tap water
(DO NOT use demineralized water);
3. Immerse the sensor in the first buffer
solution (pH 7.00 is recommended),
stir the electrode for a few seconds to
ensure that the gel layer of the pH glass
membrane is completely wetted by
buffer solution and all other fluids are
removed. Then keep the electrode still
let the measurement stabilize. Adjust
the pH reading in the pH analyser to
the value indicated on the bottle (in this
case 7.00). Go to solution 2 to do a
2-point calibration;
4. Rinse sensor thoroughly with tap water;
5. Immerse the sensor in the second buffer
solution (pH 4.00 is recommended),
stir the electrode for a few seconds
to ensure that the gel layer of the pH
glass membrane is completely wetted
by buffer solution and all other fluids
are removed. Then keep the electrode
still let the measurement stabilize. Adjust
the pH reading in the pH analyser to
the value indicated on the bottle (in this
case 4.00). Quit the manual calibration
routine by accepting the new calculated
calibration data of zero and slope.
After calibration, re-install the sensor into
the process.
Note: It is important to understand it is possible
the SC24V differential sensor does
not show the correct pH value after
buffer calibration. The reason is that
the ionic strength of the buffers is 1 mol
NaCl, which can deviate from the ionic
strength of the process. For an accurate
pH reading an extra 1 point process
calibration has to be done to compensate
for differences in ionic strength.
2.10.5 Process calibration
The following is a specific 1-point manual
calibration procedure for the process to
measure:
Set the pH analyser to “MANUAL”
calibration. Wait until the pH reading is
stable.
Adjust the pH reading in the pH analyser
to the value of the process sample that is
analysed using a laboratory measurement.
For the laboratory measurement, we
advise to a personal pH meter (e.g.
Yokogawa PH72) calibrated according to
standard (e.g. ASTM D5464, DIN 19266
or DIN 19267).
Quit the manual calibration routine by
accepting the new calculated calibration
data of zero. Slope value will be
unchanged.
m
mV
x mV
re
a
l
0
0
7
pH
id
ea
l
Fig. 2.10.5: Process Calibration
Table of Contents
TI12B00A20-01E First Edition March 2014
27
w w w.yo ko g a wa.c o m
3. Basics on ORP
3.1. Concept of ORP
Have you ever wondered what gives a sapphire its deep blue color? It comes from a
simple REDOX reaction between the titanium
(IV) and iron (II) impurities. The REDOX reaction can be seen as:
3. Basics on ORP
Ti4+ + Fe2+ → Ti3+ + Fe3+
However in order for the reaction to occur
energy must be supplied. This is easily
accomplished when ordinary white light
passes thru the crystals. The reaction
between the titanium and iron absorbs the
red, orange and yellow light regions of the
spectrum to fuel the REDOX reaction, thus
allowing only the blue light to pass thru the
crystals, resulting in the deep blue color
seen in sapphire crystals.
accept those electrons and is said to be
reduced (gains electrons).
With redox reactions we speak in terms
of the strengths of the oxidizing and reducing agents. Oxidizing agents have the
capacity or potential to acquire electrons
and become reduced. Reduction means
the gain of electrons by an atom, leading
to a decrease in the oxidation state of the
element.
Cu2+ + 2e− → Cu
While a pH value can be obtained within
seconds, a stable ORP value can take up
to several minutes, if not hours, to reach the
final equilibrium due to the type of reactions
and their reaction rates. The ORP measurement behavior is strongly influenced by the
metal surface condition. For example, a new,
unconditioned ORP electrode will show different values than an ORP electrode that has
been conditioned and considered in use.
Reducing agents donate electrons and
therefore become oxidized. Oxidation
means the loss of electrons from an atom,
leading to an increase in the oxidation state
of the element.
Fe → Fe2+ + 2e−
A simple working definition for ORP is a
solution’s capacity for electron transfer
known as oxidation or reduction, given in
millivolts. The measurement of ORP is the
reading of the voltage potential between
the measuring electrode and a reference
electrode. Depending on the solution being
measuring, the ORP electrodes will serve
as either an electron donor or an electron
acceptor. ORP is similar to pH in that pH
indicates how acidic or basic a solution is
based on the hydrogen ion activity within
the solution and ORP indicates the reductionoxidation status of a solution based on the
collective electron activity within the solution.
Since reduction and oxidation reaction
occur simultaneously, the formulas for the
two half reactions shown above, (the
reaction between iron and copper (ii) sulfate
solution) are combined and result in the
following:
Fe + CuSO4 → FeSO4 + Cu
Fig. 3.1a
The Copper, Cu, being the oxidizing agent
while the Iron, Fe, is the reducing agent.
Another example is the reaction between
hydrogen and fluorine in the process
of making hydrogen fluoride (HF). The
hydrogen (H2) is being oxidized and fluorine
(F2) is being reduced:
H2 + F2 → 2 HF
The two half-reactions are as follows:
the oxidation reaction is:
H2 → 2H+ + 2 e−
3.2. The ORP Scale
Shown in Figure 3.2 is a section of the
typical ORP scale. The full range is typically
1500 mV to -1500 mV.
Just like with pH, all ORP electrodes are
designed to produce 0 mV at pH 7. When
we look at the pH scale an acid is defined
as a substance that is capable of liberating
and the reduction reaction
is:
F2 + 2e− → 2F−
Fig. 3.1b
ORP (Oxidation-Reduction Potential) is the
measurement, in millivolts, of a solution’s
capacity for electron transfer (oxidation or
reduction). ORP measurement may also be
called REDOX for REDuction OXidation. The
name reflects that fact that within a chemical
reaction reduction and oxidation are complementary; one cannot occur without the
other. If one species undergoes oxidation
(loses electrons) then another species must
When a chemically inactive
metal electrode is placed
into a solution where an
oxidation-reduction reaction
is taking place, an electric
potential appears at the
electrode. This potential
is called the oxidationreduction potential.
M
E
Liquid being
measured
Fig. 3.1c
Table of Contents
28 TI12B00A20-01E First Edition March 2014
Indicating
electrode
(Pt or Au)
KCl solution
Junction
Reference
electrode
(Ag/AgCl)
Back to the pHuture
Soda
Tap Water
Bottled Water
Vitamin C
Green Tea
Cod Liver Oil
Living Water
– ORP (mV)
+ ORP (mV)
Fig. 3.2
An ORP system can be defined in the
same manner. Unlike pH, ORP values are
affected by all oxidizing and reducing
agents, not just acids and bases which only
influence a pH measurement. Since ORP
is the direct measurement of electrons in
transit during Oxidation-Reduction reactions,
under oxidizing conditions, the measuring
probe loses electrons to the solution, which
creates a positive potential; in a reducing
environment, electrons are donated to the
probe, producing a negative potential.
Since a reducing agent is capable of
accepting an electron and an oxidizing
agent is capable of losing an electron; it
can be said that the stronger the reducing
agent the more negative the ORP value, and
the stronger the oxidizing agent the more
positive the ORP value.
For example:
Acid Permanganate solution is strongly
oxidizing: it strongly attracts electrons
from the ReDOx electrode, so the ReDOx
potential is highly positive.
Opposite to that would be, Sulfite solutions
are strongly reducing. It pushes electrons
into the electrode, so the REDOX potential is
strongly negative.
While pH is a specific measure of the
Hydrogen ion concentration in solution, ORP
only provides relative measures of chemicals
and cannot discriminate one from another.
even though ORP is non-ion specific, it
is an inexpensive and useful method for
controlling and monitoring the activity of
various compounds such as chlorine, ozone,
bromine, cyanide, chromate, and many
others.
3.3. Measuring the ORP Value
ORP is measured in milivolts (mV), with
no correction for solution temperature.
Similar to pH, ORP is not a measurement
of concentration directly, but of activity
level. ORP is the measure of the ratio of
the activities of the oxidizing and reducing
species in a solution. The ORP value of
a particular material results in either a
positive or negative mV output; the value is
determined by the size of the atom of the
material and the number of electrons found
in its outer electron shell. The response
speed of the process measurement varies
with the concentration of the ReDOx in the
system; higher concentrations are faster and
lower concentrations are slower.
Once again, the German physical chemist
and physicist, Walther Hemann nernst
found that a potential difference occurs
between a metal object and a solution that
contains ions of the same metal, when the
object is immersed in the solution.
In electrochemistry, the Nernst equation is
an equation that can be used (in conjunction
with other information) to determine the
equilibrium reduction potential of a halfcell1) in an electrochemical cell2). It can
also be used to determine the total voltage
(electromotive force) for a system.
nernst formula:
3. BAsICs On ORP
hydrogen ions and a base is a substance
capable of absorbing hydrogen ions.
Therefore every acid has its complementary
base. When you look at the pH scale at 0
mV a solution is neutral (it is neither acidic
or alkaline), but as you move above 0 mV
the solution is considered to be acidic, and
when you move below 0 mV the solution is
considered alkaline or basic. Some common
liquids and their respective ORP values are
shown in Figure 3.2. Soda is known to have
a pH value of around 2.00; shown here the
respective ORP value for soda is approximately 400 mV. Indicating that a positive
mV reading (or below a pH 7) is associated
with the charge of the Hydrogen ion, H+,
and the solution is said to be acidic. Where
as a negative mV (or above pH 7) is associated with the charge of the Hydroxyl ion, OH–,
and the solution is said to be alkaline or basic.
E h = E0 + RT/nF log Aox/Ared
In which….
Eh =
is the Oxidation Reduction Potential
value of the reaction
is the standard potential that is
E0 =
particular to the reaction series and
that has a constant value, that not
is affected by Aox/Ared, temperature,
etc.
RT/nF =is the nernst number
Aox = is the activity of the oxidant
Ared = is the activity of the reductant
Note 1: A half cell is a structure that contains a
conductive electrode and a surrounding
conductive electrolyte separated by a
naturally-occurring Helmholtz double
layer. Chemical reactions within this layer
momentarily pump electric charges between
the electrode and the electrolyte, resulting in
a potential difference between the electrode
and the electrolyte.
Note 2: An electrochemical cell is a device capable
of either deriving electrical energy from
chemical reactions, or facilitating chemical
reactions through the introduction of
electrical energy.
Table of Contents
TI12B00A20-01E
First Edition
March 2014
29
w w w.yo ko g a wa.c o m
Element
Reduction Half-Reaction
Standard Reduction
Potential E0(V)
F2
F2 + 2 e- → 2 F-
2.870
Au
Au + 3 e →Au
1.420
Cl2
Cl2 + 2 e- → 2 Cl-
1.358
Br2
Br2 + 2 e- → 2 Br-
1.065
Hg
Hg2+ + 2 e- → Hg
0.851
Ag
Ag+ + e- → Ag
0.800
I2
I2 + 2 e- → 2 I-
0.535
Cu
Cu2+ + 2 e- → Cu
0.340
H2
2 H+ + 2 e- → H2
0.000
Pb
Pb2+ + 2 e- → Pb
-0.126
Sn
Sn2+ + 2 e- → Sn
-0.136
Ni
Ni2+ + 2 e- → Ni
-0.230
Cd
Cd + 2 e → Cd
-0.403
Fe
Fe2+ + 2 e- → Fe
-0.409
Cr
Cr3+ + 3 e- → Cr
-0.740
Zn
Zn2+ + 2 e- → Zn
-0.763
Al
Al3+ + 3 e- → Al
-1.706
Mg
Mg + 2 e → Mg
-2.375
Na
Na+ + e- → Na
-2.710
Ca
Ca2+ + 2 e- → Ca
-2.760
K
K+ + e- → Lk
-2.292
Li
Li+ + e- → Li
-3.040
Strength of Oxidizing Agent
3. BAsICs On ORP
Strongest
Oxidizing
Agent
3+
2+
2+
-
-
-
Strength of Reducing Agent
Some Standard Reduction Potentials in Aqueous Solution at 25°C
Strongest
Reducing
Agent
The reference point for all oxidation or reduction reactions, are compared to the hydrogen ion/ hydrogen (H+/H2) reaction; which has
a standard potential, E0, of 0 mV.
Table 3.3a
2H+ + 2e - → H2
Oxidized atom
free electrons
reduced atom
Tables for standard potentials, E0, as seen
in Table 3.3a, for various reactions and
their half reaction can be found in various
General Chemistry Textbook reference
materials3).
a silver/silver chloride wire in a potassium
chloride electrolyte solution. It may be either
free flowing or gel filled. The measuring
probe is typically platinum though some
other inert metals have been tested.
The tables are usually written as reduction
reactions, showing the free electrons and the
oxidized atom on the left and the reduced
atom on the right hand side of the reaction
equation.
When measuring ORP an important feature
to remember is that unlike pH, temperature
compensation is not normally used for ORP
measurements. Temperature does have
two distinct effects on ORP measurements;
however it is not compensated for because:
A typical industrial ORP measurement loop
is similar to that used for pH measurement.
It includes a high input impedance analyzer
a reference electrode, measuring electrode,
and system ground.
• Since ORP is non-ion specific
measurement, the chemistry of the redox
reaction can be quite complex, especially
if several ionic species involving varying
numbers of electrons transferred contribute
to the reaction / oxidation reduction
potential.
• Most ORP measurements are done at
constant temperatures, such as in process
measurement and control.
• The isopotential point (the point of thermal
independence) of an ORP system is only
relative to the particular redox reaction
and therefore there is no “standard”
isopoint for the overall ORP reaction.
The reference electrode is typically a
standard pH reference electrode, normally,
Table of Contents
30
TI12B00A20-01E
First Edition
March 2014
Note 3: Oxtoby, nachtrieb, Freeman. 1994.
Chemistry Science of Change.
Philadelphia: Saunders College
Publishing.
Back to the pHuture
3.4. Composition of the Measuring
Electrode
ORP/Redox is a potentiometrical
measurement of the oxidizing/reducing
power of a liquid. An ORP measuring
electrode is similar to that of a pH
measuring electrode, except it is normally
constructed of an inert (noble) metal.
The Oxygen bonds to the surface in strong
oxidizing solutions and hydrogen bonds to
the surface in strongly reducing solutions.
Some anti-corrosion chemicals added to
cooling towers and pasteurization processes
perceive the electron active surface of the
platinum as corrosion and passivates it. All
of these surface reactions can result in slow
response. Any surface coatings that insulate
the platinum surface from the solution will
decrease the speed of response.
The metallic electrode can be classified into
three distinct types of ORP electrode.
An ORP measuring electrode can either
be a separate electrode (as seen in Figure
3.4a), a combination ORP/Reference
electrode (as seen in Figure 3.4b), or a
combination ORP/pH measuring electrode
(as seen in Figure 3.4c). The choice
depends on the application as well as
customer installation requirements.
i) The first consists of a metal in
contact with a solution of the same
metal ions. i.e. a silver electrode
placed in a solution of silver nitrate,
which will develop a potential
proportional to the silver ion activity.
ii) The second consists of a metal
electrode coated with a sparingly
soluble salt of metal, in contact
with a solution containing the anion
of the metal salt. i.e. a silver-silver
chloride electrode in a potassium
chloride solution.
iii) The third kind, and most common,
consists of a noble metal in contact
with a solution containing both the
oxidized and reduced forms of an
oxidation-reduction system. This is
typically a platinum.
Note 1: C
hemisorption (or chemical
adsorption) is adsorption in which the
forces involved are valence forces
of the same kind as those operating
in the formation of chemical
compounds.
Chemisorption means to take up and
chemically bind (a substance), in this
case Oxygen, onto the surface of
another substance.
3. Basics on ORP
The most common metal used is platinum.
Platinum, which is considered the standard,
has excellent chemical resistance but suffers
slightly from Chemisorption1) of oxygen;
which slows down the measurement
response time. Meaning that the surface can
absorb organic compounds and it may be
attacked by sulfides and cyanides in strongly
reducing solutions, i.e. such as solutions with
redox potentials less than -500 mV.
The basic measurement principle is that the
measuring electrode will give up electrons
to an oxidant or accept electrons from
a reductant, without interfering with the
chemical reactions that are taking place
within the solution.
Reference system
Platinium wire
band
Glass housing
Platinum wire
Fig. 3.4a: Separate ORP Measuring
Electrode
Ag/AgCl
Wire pin
pH Glass
housing
Fig. 3.4b: Combination ORP/Reference
electrode
Ag/AgCl
Wire pin
Platinum
Fig. 3.4c: Combination ORP/pH measuring
electrode
Table of Contents
TI12B00A20-01E First Edition March 2014
31
w w w.yo ko g a wa.c o m
Et
3.5. Composition of the Reference
Electrode
ORP can be measured relative to any practical
or theoretical reference electrode such as Ag/
AgCl, or SHE (Standard Hydrogen Electrode)
like described in section 2.5.2 and 2.5.5.
3.6. The Measuring Circuit
3. Basics on ORP
An ORP probe is really a millivolt meter,
measuring very tiny voltages generated
across a circuit formed by a measuring electrode (the positive pole of the circuit normally
platinum), and a reference electrode (the
negative pole), with the process solution in
between. The difference in voltage between
the two electrodes is what is actually being
measured by the analyzer.
These voltages give us an indication of
the ability of the oxidizers or reducers
within a solution. The speed of response
varies with the concentration of the redox
system; high concentrations are fast and low
concentrations are slow.
Et
E3
E4
R elecrolyte
R elecrolyte
E2
E5
R membrane
R diaphragm
Fig. 3.6a
R
liquid
E1
E1= P otential between the ORP metal surface and the process
E2= Potential between reference electrode
and the electrolyte
E3= Potential that develops at the surface
of the electrolyte and the process
The sum total of these potential differences is
measured by the signal convertor.
E t =E1 + E2 + E3
The Figures 3.6a/b shows that following
potentials are of the most significance:
It is also important that within the analyzer
being used for detection, there must be
3.7. Standard ORP vs pH
Compensated ORP (rH)
indicate the activity of chlorine in a solution.
Since addition of chlorine increases the
oxidizing capability of water, measurement
of the ORP provides a useful indicator of
the quantity of active chlorine present. This
is very important when the chlorine is being
used as a biocide agent to control algae
growth in the process. One drawback
however, is that pH changes also affect the
oxidizing potential of the available chlorine
and the resultant ORP value.
ORP measures the ratio of the activities of the
oxidizing and reducing species in a solution.
This is a measure of the solutions ability
to oxidize or reduce another substance.
As an oxidizer is added to the process, it
“steals” electrons from the surface of the ORP
measuring electrode, causing it to become
more positively charged. Continuing to add
oxidizer generates a higher and higher
positive voltage. The role of an ORP system
is to measure these tiny voltages generated
across a circuit formed by a measuring
electrode (the positive pole of the circuit,
normally platinum), and a reference electrode
(the negative pole, reference electrode),
immersed in the solution.
ORP system are typically rugged, but do
have some limitations. For example, when
ORP is used with a chlorine-based sanitation
system, it will not indicate the chlorine
concentration in parts per million. It will
however, indicate the effectiveness of the
chlorine as an oxidizer. ORP can be used to
Fig. 3.6b
E3
E2
Metal Pin
connected
E1 to Platinum
a high impedance (resistance) in order to
measure the very tiny voltages (or charge
build up) being generated by the constant
acceptance and giving up of electrons on
the ORP electrode.
The ideal conditions described above
cannot always be completely realized in
practice. A small potential difference may
exist in the reference and is acceptable
because most mV changes measured for
ORP within solutions are large.
a large mV reading to start with, but as
it oxidizes with water reacting with the
Calcium Thiosulfite to form a salt, the mV
reading would decrease.
Below a value of 1.9 pH, chlorine exists as
a diatomic molecule (Cl2) in water. As the
pH increases above 1.9, chlorine oxidizes
water to produce HOCl and the ORP
millivolt reading will go down. As the pH
value continues to increase HOCl further
dissociates into OCl- above a pH of 7.3.
Since ORP varies with pH changes, as
well as changes in the chlorine levels, we
must compensate for the effects of any pH
changes. This can be done by measuring
the pH and ORP independent of one
another and then calculating the effect of
the pH change on the ORP using formulas
and graphs.
A simpler and more direct method is
to compensate for the pH changes by
replacing the standard Ag/AgCl reference
electrode normally used with a pH
measuring electrode. This is known as pH
Compensated ORP (rH).
HOCl being more active that OCl- has a
higher ORP value. So, as the pH increases,
an ORP sensor detects a decrease in value
which reflecting the decrease in HOCl.
Therefore, if we were using ORP to monitor
the diatomic Cl2 level, we would have
Since the pH measuring electrode output
changes as the pH of the process changes
it acts as a moving reference effectively
cancelling out any change in pH and
leaving only the mV value which is due to
changes in diatomic chlorine (Cl2) levels.
Table of Contents
32 TI12B00A20-01E First Edition March 2014
Back to the pHuture
3.8. Standard Maintenance and
Calibration
Maintenance and calibration for an ORP
system has similarities with those methods
and procedures used for both pH and
conductivity systems. The problems and
maintenance procedures associated with the
ORP reference electrode, for example, are
handled the same as they are for a pH loop.
The maintenance of the ORP measuring
electrode however, is handled in much the
same way as a conventional conductivity
sensor. Calibration of the ORP system is
also similar to how a conductivity loop is
calibrated.
If the various forms of chemical cleaning
are not sufficient to achieve an accurate
measurement and response time, a last resort
would be to polish the platinum surface with
a 600 grit wet-dry emery cloth or a 1-3
micron alumina polishing powder to remove
any surface pitting or stubborn coatings.
ORP electrodes should need less frequent
calibration than a typical pH sensor since
the redox potential is a characteristic of the
interaction between the platinum and the
redox equilibrium. However, it is prudent to
periodically verify the performance of the
measuring system by placing the electrode
in a solution with a known potential and
calibrate is needed to correct for the
reference side of the sensor.
Standard ORP Solutions
When verification or calibration of an ORP
sensor is required, there are two types of
Standard Solutions that are commonly used.
The first are premade solutions designed to
provide a specific stable mV value, typically
one that falls within the process ORP range.
The second type of solutions, and probably
the most common, are those that are made
using the standard pH 4 and pH 7 buffers
with quinhydrone crystals mixed in until
saturation is reached. Either of these pH
buffer solutions can be used for calibration
of an ORP measuring system and are
very practical if pH loops are also being
maintained. Preparation and use of both
types of solutions are discussed below:
Quinhydrone1) Solution
To create an ORP solution using a pH buffer
(either 4.0 or 7.0) stir in a small amount,
approximately < 0.5 gm, of quinhydrone
into 200 mls of solution. Quinhydrone is
not very soluble, so only a small amount
will dissolve in the buffer changing the
solution to an amber color. If all of the
quinhydrone does dissolve, then continue
to add small amounts and stir again.
Saturation is achieved when a small amount
of quinhydrone remains un-dissolved after
mixing.
Whether it is a 4.0 or a 7.0 buffer you are
using, Table 3.8a shows the mV reading you
should obtain depending on which reference
electrode is being used. As an example, a
quinhydrone/pH 4.0 solution should give a
253 mV (± 30 mV) at 25°C for a reference
electrode that has 3M KCl internal fill.
3. Basics on ORP
Cleaning the measuring electrode will
improve accuracy and the sensors response
time. The frequency of maintenance,
which includes cleaning and calibration, is
determined by how the process affects the
electrode. Methanol, isopropyl alcohol or
a detergent can be used to remove oily or
organic coatings while those applications
where removal of any anti-corrosion
chemicals or mineral deposits, soaking the
electrode in 10% nitric acid for about 10-15
minutes is a good starting point.
After any cleaning procedure, the electrode
should be allowed to soak in clean tap
water for at least 30 minutes to allow
residual chemicals to dissipate and the
sensor to recover. After calibration, when the
electrodes are placed back in the process,
they should be allowed to equilibrate for at
least 15 minutes. For optimal operation, the
sensors should be installed in an area with
good agitation as process flows past the
electrode helps to keep the platinum sensing
element clean.
Note 1: The quinhydrone powder poses a
moderate health risk, causing irritation
of the lungs with prolonged exposure.
The premade calibration solutions
are fairly innocuous unless ingested
in large amounts. Both types should
be handles carefully following good
laboratory practices.
Note 2: SCE = Saturated Calomel Electrode
Note 3: SHE = Standard Hydrogen Electrode
Table 3.8a mV value of ORP solution made with pH Buffers and Quinhydrone
ORP Value (mV)
Reference
Electrode
pH 4 Buffer Solution
pH 7 Buffer Solution
20°C
68°F
25°C
77°F
30°C
86°F
20°C
68°F
25°C
77°F
30°C
86°F
Ag/AgCl
(1M KCl)
236
231
226
61
54
47
Ag/AgCl
(3M KCl)
257
253
249
82
76
Ag/AgCl
(sat. KCl)
268
263
258
92
86
79
Calomel
(sat. KCl)
---
218
---
---
41
---
SCE2)
223
218
213
47
41
34
SHE3)
470
462
454
295
285
275
70
Table 3.8a
Table of Contents
TI12B00A20-01E First Edition March 2014
33
w w w.yo ko g a wa.c o m
Pre-Made Stabilized ORP Solutions
chart below, you have to know what (1)
reference solution is used in the reference
electrode and (2) what reference solution the
known premade solution is being compared
to. For example, if you have a premade
250 mV solution that is referenced to SHE
(Standard Hydrogen Electrode) and the
reference electrode in your measuring loop
uses a 1 M KCl fill solution, then on the
transmitter you would NOT read 250 mV,
but instead you would read only 19 mV
at 25° C. This is the 250 mV value on the
Reference electrodes with different internal
fill solutions will have different mV outputs
when they are put in the same Standard
Solution. This is because the Standard
Solution was prepared with one specific
reference fill solution in mind. Table 9.2
lists in the left-most column, some of the
most commonly used reference electrode fill
solutions. Across the top of the table are the
possible reference fill solutions that Standard
Solution was prepared against. To use the
solution minus the 231 mV value shown as
the difference between the SHE and the 1M
KCl references. This would be 19 mV.
Note 1: SCE = Saturated Calomel Electrode
Note 2: SHE = Standard Hydrogen Electrode
Table 3.8b mV offset between Various Reference Electrode Solutions
ToAg/AgCI)
(3MKCI)
3. Basics on ORP
To SCE1)
ToAg/AgCI
(sat. KCI)
ToAg/AgCI)
1M KCI)
To SHE2)
20°C
68°F
25°C
77°F
30°C
86°F
20°C
68°F
25°C
77°F
30°C
86°F
20°C
68°F
25°C
77°F
30°C
86°F
20°C
68°F
25°C
77°F
30°C
86°F
30°C
86°F
30°C
86°F
30°C
86°F
From SCE1)
–
–
–
+34
+35
+36
+45
+45
+46
+13
+14
+16
+241
+241
+241
From Ag/AgCI
(3M KCL)
-34
-35
-36
–
–
–
+11
+10
+9
-21
-22
-23
+205
+205
+205
FromAg/AgCI
(sat KCL)
-45
-45
-45
-11
-10
-9
–
–
–
-32
-31
-30
+202
+199
+196
From Ag/AgCI
(1M KCL)
-13
-14
-16
+21
+22
+23
+32
+31
+30
–
–
–
+234
+231
+228
From SHE2)
-247
-244
-241
-213
-209
-205
-202
-199
-196
-234
-231
-228
–
–
–
Table 3.8b
Stir in a small amount of approximately
< 0.5 gm, of quinhydrone into 200 mls of a
pH buffer solution. Quinhydrone is not very
soluble, only a small amount will dissolve in
the buffer changing the solution to an amber
color.
When verification or calibration of an ORP
sensor is required, there are two types of
Standard Solutions that are commonly used.
The first are premade solutions designed to
provide a specific stable mV value, typically
one that falls within the process ORP range.
The second type of solutions, and probably
the most common, are those that are made
using the standard pH 4 and pH 7 buffers
with quinhydrone crystals mixed in until
saturation is reached. Either of these pH
buffer solutions can be used for calibration
of an ORP measuring system and are
very practical if pH loops are also being
maintained. Preparation and use of both
types of solutions are discussed below:
Proper calibration
C. Allow stabilization time then adjust
to the correct value.
The following steps are commonly used for
calibration of an ORP loop.
1. C
lean the ORP & Reference electrodes
FIRST.
2. M
ake FRESH Buffer Solutions with either
Quinhydrone Crystals (See section 3.8),
or pour a fresh sample of a premade
stabilized ORP solution.
A. ORP Buffers are best made just
before using.
B. Never keep (store) ORP calibration
solutions.
4. Rinse the electrode between calibration
measurements if checking at a second buffer
value.
If a short span is found (less than a +150
mV change between the first solution and
the second solution), the platinum/ gold
measuring surface may be coated and
the electrode should be re-cleaned and
re-calibrated.
3. Perform a single (1) point calibration.
A. Pick a solution with a mV value
closest to the control point.
B. If possible adjust (heat) the solution
to one of the temperatures shown in
the mV tables above.
Table of Contents
34 TI12B00A20-01E First Edition March 2014
Back to the pHuture
Notes
3. Basics on ORP
Table of Contents
TI12B00A20-01E First Edition March 2014
35
w w w.yo ko g a wa.c o m
4. PRODUCTs aND feaTURes
Note: not all products are offered globally
please contact local office.
No. of sensor inputs: Single or dual sensor
measurement.
The purpose of this section is to share the
benefits which a customer can get when he
uses Cutting edge Technology of Yokogawa.
It includes information on electronics
(converter/transmitter), electrodes, and their
features and functionalities.
YOKOGAWA
Made in the Netherlands
YOKOGAWA
4.1 electronics/ Transmitters/
Converters
A) 4 – wire type analyzer:
(Yokogawa model: PH450G series)
Fig. 4.01 Control Function
Impedance monitoring: For both pH and
reference cell available (fig. 4.02, 4.03). In
addition to this a dynamic sensor checking
coupled with the wash cycle function assure
trouble free and accurate analysis with a
minimum of maintenance.
4.PRODUCT & FeATURes
Fig. 4.03 Impedance monitoring
Human Machine Interface: High resolution
graphical display and the touchscreen operation that make all information visible to the
operator. Configuration with the touchscreen
as easy as operating a Tablet P.C. (fig. 4.04,
4.05). Choose the language of choice (min.8
languages available to choose) and on screen
instructions assure that the best configuration
for the application be obtained.
A-B
E8
Type: 4 wire type
Power supply: Universal AC voltage – Accept
range; 90 to 264 V AC Ratings; 50/60 Hz,
Power Consumption; 15 VA OR Universal DC
voltage accept range 10.8 to 26.4 V DC,
Power Consumption; 10 W
A
B
E7
E6
E5 E4
E1 E2 E3
Features: Advanced temperature
compensation functionality, preloaded
calibration standards and stability checks.
True multivariable analyzer: Combine
pH with Temperature and ORP (Redox)
measurement and all these measurements
can be utilised through the different output
functions: two mA current outputs, four
independent SPDT contact outputs and
HART®.
Full functionality: PID control on either mA
output(s) or on contact output(s) and with
integral wash function available in analyser
(fig. 4.01).
Fig. 4.02 Stability noise rejection
Table of Contents
36
TI12B00A20-01E
First Edition
March 2014
Fig. 4.04, 4.05 Human machine interface
Back to the pHuture
Mounting: IP66/NEMA 4X 1/2DIN enclosure
for field mounting and panel mounting
Trending: Trending display available up to
2 weeks
Data storing: On-screen logbooks store
calibration data, configuration changes and
events
Display: Main screen displays containing
primary variable in large font (user
selectable), other process variable(s) in
small font, Unit symbols, Tag number (user
programmable), Process description (user
programmable), Status of alarm output(s),
Status indicator during HOLD and WASH
situation, Main function keys.
B) 2 – Wire type analyzer:
(Yokogawa model: FLXA21 series)
Input ranges: Analyser accepts ranges such
as pH: -2 to 16 pH, ORP: -1500 to 1500
mV, rH: 0 to 100 rH and Temperature
- Pt1000: -30 to 140ºC, Pt100: -30 to
140ºC, 350Ω (DKK): -30 to 140ºC, 5k1:
-30 to 140ºC, 6k8: -30 to 140ºC, PTC10k:
-30 to 140ºC, NTC 8k55: -10 to 120ºC,
3kBalco: -30 to 140ºC.
Accuracy: pH input: ≤ 0.01 pH, ORP input:
≤ 1 mV, Temperature: ≤ 0.3ºC (≤ 0.4ºC for
Pt100), Response time: Step response is less
than 4 sec for 90% (pH 7 - pH 4)
Safety Standards: In line with EN 61010-1
CSA C22.2 No.61010-1, UL 61010-1
Fig. 4.07 FLXA21 inside
Input circuitry: Input circuitry of analyser
is a dual-amplifier system with solution
ground. Measuring electrode and reference
electrode inputs amplified separately against
the solution ground contact. Following
a differential amplifier, normal signal
processing takes place. This configuration
provides the best immunity to noise,
stray solution potentials and earth loops
(fig. 4.02).
Ambient conditions: Handle temperature
between -20 to +55ºC (-5 - 130ºF) and
Storage temperature -30 to +70ºC (-20 160ºF). Humidity 10 to 90% RH at 40ºC
(100ºF) (non-condensing).
No. of Sensor inputs: Single or dual sensor
measurement
Power supply: Nominal 24 V DC loop
powered system
HMI: Easy touch screen operation and
simple menu structure in 12 languages.
Menus of display, execution and setting
displayed in a selected language.
Immediate measurement: For immediate
measurement, analyser has offer quick setup
functionality. The quick setup screen appears
when the analyzer is powered.
Features: Analyser have sensor diagnostics
and sensor wellness indication which make
measurement reliable.
Data protection: EEPROM for configuration
data and logbook and Lithium cell for clock.
Watchdog timer checks microprocessor.
(fig. 4.06)
4.Product & Features
EMC standards: EN 61326-1 Class A, Table
2 (For use in industrial locations) EN 613262-3 EN 61000-3-2 Class A EN 61000-3-3
Korea Electromagnetic Conformity Standard
Type: 2 wire type
Data store: Logbook of events and
diagnostic data available with useful
information source for maintenance.
Fig. 4.08 Prediction
Advanced temperature compensation:
Analyser offers three additional modes for
advanced temperature compensation viz;
Full compensation for strong acids and
alkali’s using NEN6411 algorithm, linear
temperature coefficient setting, and Matrix
temperature compensation.
Suitability to industrial environment:
Analyser is designed with the enclosure of
polycarbonat or stainless steel or stainless
steel with corrosion resistant coating.
Fig. 4.06 Watchdog timer checks
Table of Contents
TI12B00A20-01E First Edition March 2014
37
w w w.yo ko g a wa.c o m
Sensor 1
Sensor 1
Fails
Repair Sensor 1
Sensor 2
Manually
Switch Back to Sensor 1
Automatically
Switch to Sensor 2
Alive
Alive
mA output
Alive
Dead
Alive
mA output
Alive
mA output
Alive
Fig. 4.09 Sensors check
Output Signal: Bi-directional HART digital
communication superimposed on mA
(4-20mA) signal
Ambient conditions: Temperature between
-20 to +55ºC (-5 - 130ºF) and Storage
temperature -30 to +70ºC (-20 - 160ºF).
Humidity is 10 to 95% RH at 40ºC (100ºF)
(non-condensing).
4.Product & Features
Safety Standards: In line with EN61010-1,
UL 61010-1
EMC standards: EN61326-1 Class A,
Table 2 (For use in industrial locations),
EN61326-2-3, AS/NZS CISPR11, Korea
Electromagnetic Conformity Standard
ADDITIONAL POINTS DUAL SENSOR
MEASUREMENTS:
Additional functionalities: Offers calculated
data function and redundant system.
Redundant system: On the redundant
system built on two measuring parameters
of two sensor inputs, main output parameter
automatically switched over to the second
sensor output in case of the main sensor’s
failure condition (fig. 4.09).
HAZARDOUS AREA LOCATION:
Applicable standard:
a) ATEX Intrinsically safe approval
Applicable standards: Electrical
Apparatus for Potentially Explosive
Atmospheres in conformity with EN
60079-0:2009 General requirements,
EN 60079-11:2007 Intrinsic safety
“i”, EN 60079-26:2007 Equipment
with equipment protection level (EPL)
Ga, EN 60529:1992 Degrees of
protection provided by enclosures (IP
Code) Type of protection II 1G Ex ia IIC
Ga, Group: II, Category: 1G, T4: for
ambient temperature:–20 to 55ºC, T6:
for ambient temperature:–20 to 40ºC,
Atmosphere pressure: 80kPa (0.8bar) to
110kPa (1.1bar), Degree of Protection of
the Enclosure: IP66
T4: for ambient temperature:
-20 to 55°C
T6: for ambient temperature:
-20 to 40°C
Atmosphere pressure: 80 kPa (0.8 bar)
to 110 kPa (1.1 bar)
Degree of Protection of the Enclosure:
NEMA Type 4X and IP66
d) CSA
b) IEC
Ex intrinsically safe approval
Applicable standards: IEC 60079-0:
2007 Part 0: General requirements, IEC
60079-11: 2006 Part 11: Intrinsic safety
“i”, IEC 60079-26: 2006 Part 26:
Construction, test and marking of Group
II Zone 0 electrical apparatus IEC
60529: 2001 Degrees of protection
provided by enclosures (IP Code), Type
of protection Ex ia IIC Ga, T4: for
ambient temperature:–20 to 55ºC, T6:
for ambient temperature:–20 to 40ºC,
Atmosphere pressure: 80kPa (0.8bar) to
110kPa (1.1bar), Degree of Protection of
the Enclosure: IP66
c) FM
Intrinsically safe and nonincentive
approval:
Type of protection:
Class I, Division 1, Groups A, B, C and
D (Intrinsically Safe)
Class I, Division 2, Groups A, B, C and
D (Nonincendive)
Class I, Zone 0, in Hazardous
(Classified) Locations (Intrinsically Safe)
Class I, Zone 2, Group IIC, in
Hazardous (Classified) Locations
(Nonincendive)
AEx ia IIC
For all protection type,
Table of Contents
38 TI12B00A20-01E First Edition March 2014
Intrinsically safe and nonincentive
approval:
Type of protection:
Class I, Division 1, Groups A, B, C and
D (Intrinsically Safe)
Class I, Division 2, Groups A, B, C and
D (Nonincendive)
For all protection type,
T4: for ambient temperature:
-20 to 55°C
T6: for ambient temperature:
º-20 to 40°C
Atmosphere pressure: 80 kPa (0.8 bar)
to 110 kPa (1.1 bar)
Ambient Humidity: 0 to 100% (No
condensation)
Degree of Protection of the Enclosure:
Type 4X
Type of protection (E60079)
Ex ia IIC
T4: for ambient temperature:
-20 to 55°C
T6: for ambient temperature:
-20 to 40°C
Atmosphere pressure: 80 kPa (0.8 bar)
to 110 kPa (1.1 bar)
Ambient Humidity: 0 to 100% (No
condensation)
Degree of Protection of the Enclosure: IP66
Back to the pHuture
4.2. sensors/ electrodes
2) Combined pH/ORP sensor for
harsh applications with fluctuating
pressure and/or temperature
3) Combined pH/ORP sensor
in applications where chemical
resistance is a vital issue:
(Yokogawa model: FU20 series)
(example: pure water applications):
(Yokogawa model: FU24 series)
(Yokogawa model: PH20 series)
Reference system: Long life
saturated Ag/AgCl reference
system.
Reference system: Long life
saturated Ag/AgCl reference
system
Reference junction: PTFe
reference diaphragm to
prevent fouling and reduce
measurement error. Double
junction combined with iontrap to prolong the life of
the reference probe even
in chemically unfavourable
environments.
Reference junction: PTFe
reference diaphragm to
prevent fouling and reduce
measurement error. Double
junction combined with iontrap to prolong the life of
the reference probe even
in chemically unfavourable
environments
Temperature compensation:
Integral Pt1000 element
for accurate temperature
compensation.
Temperature compensation:
Integral Pt1000 element
for accurate temperature
compensation.
Additional measurement:
equipped with platinum ORP/
Le electrode for accurate
simultaneous pH- and ORP
measurements.
Additional measurement:
equipped with platinum
ORP/Le electrode for
accurate simultaneous pHand ORP measurements.
Electrolyte: Polymerized electrolyte to
extend the sensors life time.
Electrolyte: Polymerized
electrolyte to extend the
sensors life time.
Reference system: Long life
saturated Ag/AgCl reference
system
Reference junction: PTFe
reference diaphragm to
prevent fouling and reduce
measurement error. Double
junction combined with iontrap to prolong the life of
the reference probe even
in chemically unfavourable
environments
Temperature compensation:
Integral Pt1000 element
for accurate temperature
compensation.
Additional measurement:
equipped with platinum ORP/Le electrode
for accurate simultaneous pH- and ORP
measurements.
Electrolyte: Polymerized electrolyte to
extend the sensors life time.
Construction: Wide body sensors (26 mm
diameter), holds four separate measuring
elements in one unbreakable and chemical
resistant RytonTM body.
Operating range:
Designed for the ranges; pH: 0 – 14,
Temperature: -10 to 105ºC (14 to 212 °F),
Pressure: 0 to 10 bar (0 to 142 PSIG),
Conductivity: > 50 µS/cm,
Storage temp.: -10 to 50ºC (-22 to 122 °F).
Construction: The sensor body is made with
a chemical resistant RytonTM body. By using
the Bellow system integrated in the electrode,
a strong pressure compensation mechanism
created.
The build-in bellow ensures immediate
interior pressure equalization to the outside
pressure, making the sensor virtually
insensitive to external pressure variations.
A slight overpressure caused by the bellow
tension prevent fluid ingress and maintains a
positive ion flow out of the sensor.
Operating range:
Designed for the ranges; pH: 0 – 14,
Temperature: -10 to 105ºC (14 to 212°F),
Pressure: 0 to 10 bar (0 to 142 PSIG),
Storage temp.: -15 to 50ºC (-22 to 122°F)
Construction: The sensor body is made
of chemically resistant PVDF. Sensor
compensation for changes in the process
temperature and pressure.
4.PRODUCT & FeATURes
1) Combined pH/ORP sensor for
general purpose applications:
The compensation panels flex to
accommodate changes in the avoiding large
differential pressures across the diaphragm
and prevents most problems associated with
the reference junction.
Operating range:
Designed for the ranges; pH: 0 – 14,
Temperature: -10 to 105ºC (14 to 212 °F),
Pressure: 0 to 10 bar (0 to 142 PSIG),
Conductivity: > 50 µS/cm,
Storage temp.: -15 to 50ºC (-22 to 122 °F)
Table of Contents
TI12B00A20-01E
First Edition
March 2014
39
w w w.yo ko g a wa.c o m
4) pH or ORP sensor:
(Yokogawa model: PH87/97 series)
Sensor type: Retractable
Reference system: Long
life saturated Ag/AgCl
reference system with the
option of a solid wood
state reference. (Stacked
KCl impregnated wood
dicks)
Reference junction:
Coaxial porous PTFe
reference diaphragm to
prevent fouling and reduce
measurement error.
4.PRODUCT & FeATURes
Temperature
compensation: Integral
Pt1000 element for
accurate temperature
measurements.
Electrolyte: KCl electrolyte
to extend the sensors life
time.
Construction: The combination sensor
(PH97) has wide body sensor and hold
three separate measuring elements in one
unbreakable and chemical resistant PPS
40GF (RytonTM) body. The retractable
housing (PH97) be constructed with a
titanium sheath and 316SS compression
fitting.
Replacement made easy: The pH glass
electrode of a pH sensor, the platinum
or gold electrode of a ORP
sensor and junction can be
individually replaced.
a)KCl Refillable Type Sensor
Model: PH8ERP
b) KCl Filling Type Sensor Model: PH8EFP
Measuring range: pH 0 to 14
Temperature: Sensor is suitable
for temperature –5 to 105
deg C
Pressure: Sensor handle
pressure from atmospheric
pressure to 10kPa normally
and up to 500 kPa when using
medium pressure KCl reserve
tank
Temperature compensation
sensor: Pt1000
Measuring range: 2 to 12 pH
Measuring temperature:
–5 to 80 deg C
Measuring pressure:
Atmospheric pressure to 50kPa
Wetted part materials: Body;
Ryton (PPS resin), glass, titanium
or Hastelloy C, ceramics,
teflon, fluorocarbon rubber or
Daielperfrow rubber Cable;
Chlorinated polyethylene
rubber (Cable sheath)
Temperature compensation
sensor: Pt1000
Wetted part materials: Body
Ryton (PPS resin), glass, titanium
or Hastelloy C, ceramics,
fluorocarbon rubber or
Daielperfrow rubber.
KCl tube is made of Heat-resistant soft PVC
6) Combined pH Sensor for High
Purity Water:
Temperature compensation: Compensates
for the effect of fluid temperature.
(Yokogawa model: PH8EHP)
Design of holder: The dedicated holder
provides solutions to problems that arise
when measuring high-purity water.
Operating range: Designed for the ranges;
pH: 0 – 14, ORP: -5000 – 5000 mV,
Temperature: 0 to 105ºC (0 to 221°F),
Pressure: 0 to 7 bar (0 to 100 PSIG), Flow:
3 m per second
5) Ryton pH/ORP Sensors:
(Yokogawa model: PH8EFP, PH8ERP,
OR8EFG, OR8ERG series)
Sensor body made of Ryton, a strong
engineering plastic, which be with good
corrosion resistance and heat resistance
Design: The integrated-sensor design simplify
calibration with standard solutions and
maintenance.
Fig. 4.10 Combined pH Sensor for High Purity water
Table of Contents
40
TI12B00A20-01E
First Edition
March 2014
Back to the pHuture
7) Combined pH sensor in a 12 mm
design with an external solution ground:
volume. This increases the sensor lifetime
significantly compared with pH electrodes.
Conductivity: Measuring in applications
from 10 µS/cm
(Yokogawa model: SC25V series)
Material of solution ground: Titanium
Connector: Variopin connector
Temperature sensor: Pt1000 integration
in pH compartment giving highly accurate
temperature compensation
Applications:
• SC25V-ALP25 for chemically harsh
applications and high temperatures
• SC25V-AGP25 for all General Purpose
applications
sC25v
Sensor diameter and design: 12 mm.
Certification: ATex II1G ex ia IIC T3...T6 Ga
Design: Sensor design with an external
Solution ground. The design made it
possible to create a large electrolyte
CIP: CIP and Steam cleaning possible
Specifications Combined pH electrodes with Temperature + Solution ground
Type
Membrane
Resistance pH-range
in MΩ/25ºC
SC25V-AGP25 Universal pH-glass bulb
175-275
0-14
SC25V-ALP25 Chem. Res. pH-glass dome 500-700
0-14
8) Combined electrode:
Temp.
range (ºC)
-10 - 80
+15 - 130
Pressure
range kPa
0-1000
0-1000
Reference
liquid
Oversaturated KCl
Oversaturated KCl
a) SC21-AGP24
• Reference system: Ag/AgCl wire
• Design: pH bulb with cage protection
• Electrolyte & Junction: Less maintenance
due to the gelled electrolyte - Thickened
electrolyte (3.3 m.) and porous PTFe.
(Yokogawa model: SC21 series)
sC21-AAP26
sC21-ALP26
sC21-AgP26
c) SC21(C)-AGP26
• Reference system: High quality Ag/
AgCl reference system (pin) which can
stand high temperatures and temperature
fluctuations.
sC21C-AgP64
Atex
Yes
Yes
• Reference system: Double junction
(thickened saturated KCl-solution). The
built-in salt bridge prevents poisoning of
the reference system. A large area PTFe
junction to resist fouling.
• Glass design: Heavy duty glass
membrane for prolonged operation
in corrosive, abrasive and fouling
environments (withstanding traces of HF).
d) SC21(C)-ALP26
• Design: Chemical resistant, steamsterilisable pH-glass.
e) SC21C-AGP64
• Reference system: Ag/AgCl wire reference
system
• Electrolyte and junction: Less maintenance
by the combination of gelled electrolyte
-3.3 m KCl electrolyte and porous PTFe.
4.PRODUCT & FeATURes
b) SC21-AAP26
• Reference system: High quality Ag/AgCl
reference system (pin) which can stand high
temperatures and temperature fluctuations.
• Reference system: Built-in salt bridge to
prevent poisoning of the reference system.
• A large area PTFE junction to resist fouling
to a high degree.
Diaphragm
Ref.
system
Ag/AgCl (wire) PTFE
Ag/AgCl (wire) PTFE
Specifications Combined pH Electrodes (non-flow)
Type
Membrane
Resistance
pH-range Temp.
in MΩ/25ºC
SC21-AGP24
Universal
50 - 100
0 - 14
Pressure
range kPa liquid
0 - 80
1-500
pH-glass bulb
SC21(D)-AAP26 Chem. res. pH-glass bulb
250 - 400
0 - 14
0 - 110
1-500
steam-sterilisable / bulb
3
SC21-ALP26
Chem. res.
Universal
500 - 900
0 - 14
10 - 120
1-500
Universal
Diaphragm
Flow
system
3.3 m. KCI
Ag/AgCl (wire)
Thickened
Silver-silverchloride
Oversaturated Ag/AgCl (wire)
Porous PTFE 0
Porous PTFE 0
Oversaturated Ag/AgCl (wire)
Porous PTFE 0
KCI thickened Silver-silverchloride
120 - 200
0 - 14
-10 - 100
1-500
pH-glass bulb
SC21-AGP64
Reference
KCI thickened Silver-silverchloride
4
pH-glass dome
SC21-AGP26
Reference
range (ºC)
Oversaturated Ag/AgCl (wire)
Porous PTFE 0
KCI thickened Silver-silverchloride
50 - 100
0 - 14
0 - 80
1-500
pH-glass bulb
3.3 m. KCI
Ag/AgCl (wire)
Thickened
Silver-silverchloride
Porous PTFE 0
Table of Contents
TI12B00A20-01E
First Edition
March 2014
41
w w w.yo ko g a wa.c o m
9) Combined flow type pH
electrode:
Reference system: Flowing reference system
for pollution resistance, and highly stable
reference potential.
“general purpose” glass. High resistance glass
for chemical resistance. G-glass sensors for all
general applications.
Electrode connection: PG13.5 standard DIn
electrode connection.
“L” glass: The application of “L” glass is for
measurements in alkaline media with high
process temperatures. Low resistance glass for
fast response and sensitivity. L-glass sensors for
high temperature chemically harsh applications.
(Yokogawa model: SC21C-AGC55)
sC21C-AgC55
Type of sensor: Flowing electrolyte sensor
uses a ceramic reference junction
Reservoir: Use in combination with the
presurisable electrolyte reservoir to obtain a
positive flow towards the process
Design: Dimensions and design meet the
requirements of DIn 19263.
Isothermal point of intersection: pH 7 (nominal
value at 0 mV).
Maximum pressure: 1000 kPa (10 bar).
Thickness of the glass membrane
Shockproof: Bulb membrane for general
purpose. A bulb shaped glass membrane
suitable for general purpose
Heavy duty: This dome shaped glass
membrane is suitable for harsh and abrasive
applications.
10) Single glass electrode:
Electrolyte: 3.3 molal KCl
Application: For tough application where
pollution of the reference system is to be
expected. Low ionic application where
the possitive flow of electrolyte provides
the conductivity needed to measure
pH(< 50 µS/cm)
Glass type: Heavy duty pH sensitive glass
(Yokogawa model: SM21 series)
Two types of glass are available, as
described here:
”G” glass: This is used for the membranes
of electrodes in processes where the
nominal pH value varies between pH0
and pH14. Since this type of glass has a
wide application range it is also termed
4.PRODUCT & FeATURes
Selection guide separate electrodes: Glass sensors
modelcode
Bulb type
SM21(D)-AG2
SM21(D)-AG4
SM21(D)-AG6
SM21(D)-AL4
SM21(D)-AL6
Universal
shockproof
heavy duty
shockproof
heavy duty
Glass resistance
(Mohm)
25-50
50-100
120-200
300-450
600-900
T range
(°C)
0-80
0-100
0-100
15-130
25-130
Pressure
(bar)
10
10
10
10
10
Applications
remark
light, laboratory
general
Harsh
High Temperature
High T + abbresive
fast response, high sensitivity
Universal
mechanically strong
T cont. ≥ 70°C, high chem. Resistance
T cont. ≥ 70°C, high chem. Resistance
* all pH glass sensors have a sodium error of pH 0.17 above pH=13 this because the glass then will see the Na+ ion as H+ ion.
The pH reading is lower than it is in reality.
11) Single reference electrode
(Yokogawa model: SR20 series)
The selection of the correct type of junction
of a reference electrode depends on
the process conditions under which this
electrode has to function. The following
junction types are available:
1. Ceramic junction.
2. PTFe junction.
3. Glass sleeve capillary element.
Reference electrode
• Flowing type sensors for dirty applications
or (Ultra) pure water applications (often
needs a electrolyte reservoir)
• Non-flow type sensors for all general
applications
• Ceramic and PTFe junction: electrolyte flows
slowly into the process. PTFe resists dirt
Ceramic 1
PTFE 2
• Glass sleeve: very dirty applications or for
(Ultra) pure water applications because of
the larger flow-rate.
components that poison the reference
system. The gel-type electrodes have a large
area of porous PTFe junction for optimal
resistance against electrode pollution.
Non-flow type
A non-flow type reference electrode can
be used for processes that don’t contain
Table of Contents
42
TI12B00A20-01E
First Edition
Sleeve 3
March 2014
Back to the pHuture
The SR20-AC22 electrode has a flexible
PTFE tube. Therefore this electrode can
be applied in processes with frequent
temperature- and pressure fluctuations.
SR20-AC22
G 1/4"
G 1/4"
Ø 17
Features
• Easy maintenance.
• No reference liquid wastage.
•M
aximum process pressure: 1000 kPa (10 bar).
• High quality Ag/AgCl reference system
(pin) which can stand high temperatures
A) Additional features of types SR20(D)AC22
• Temperature / pressure variation
compensation.
• To be used in non-polluting fluids.
• Saturated KCl-solution (pellets).
• For low ionic applications and high
temperatures.
• Temperature range: 0 to 120ºC.
• Diaphragm resistance (25ºC) <5kΩ.
SR20-AP26
SR20-AP24
23
G 1/4"
Ø 17
23
Ø 17
23
43
Ø 12
Ø 12
Ø 12
120
120
120
Flexible
PTFE tube
3
13
Ø5
Ø 8,5
Two-phase electrode
Gel-type electrode
Double junction gel-type
Specifications Single Reference Electrodes (non-flow)
Type
SR20(D)-AC22
SR20(D)-AP24*
SR20(D)-AP26*
Temperature
range (ºC)
0 - 120
0 - 80
0 - 80
Pressure
(bar)
10
10
10
Reference
Reference system
Diaphragm
liquid
Silver-silverchloride
Saturated KCl-solution (pellets) Ag/AgCl (pin)
Ceramic
Thickened KCl (3.3 m.)
Ag/AgCl (pin)
PTFE
Thickened KCl or KNO (3.3 m.) Ag/AgCl (pin)
PTFE
3
Diaphragm
resistance/25ºC
< 5 kOhm
< 5 kOhm
< 5 kOhm
B) Additional characteristics of type
SR20(D)-AP24
• General purpose PTFE diaphragm
electrode.
• Large PTFE diaphragm.
C) Additional characteristics of type
SR20(D)-AP26
• When KCl is prohibited in the application.
• Double junction, thickened KNO3 in
buffer compartment.
• Large PTFE diaphragm against pollution.
• For applications where KCl cannot be used
Flow Type Ref. Electrodes
Pressure compensated Reference Electrode
In processes with pressure variations, the
composition of the electrolyte may change
as a result of process liquid penetration into
the electrode. Any change in composition of
the electrolyte may cause a measuring error
or even poisoning of the reference system
of the electrode. To alleviate this problem,
the electrode with an integral pressure
compensation system (“Bellomatic” -type
electrode”) can be the solution.
Specifications Single Reference Electrodes (flow)
Type
Temp.
Pressure
range (ºC) range
SR20(D)-AS52 0 - 100
Atmospheric
0 - 1 MPa
SR20(D)-AC32 0 - 120
Reference
liquid
KCl-solution
(3.3 m.)
KCl-solution
(3.3 m.)*
Features
• Liquid flow output preventing diaphragm
fouling and poisoning the reference system.
• High quality Ag/AgCl reference system
(pin) which can stand high temperatures
and temperature fluctuations.
• Standard 3.3 m. KCl electrolyte, at temp.
o
above 70 C thickened electrolyte is
advised.
• Automatic compensation for process
pressure variations.
• Chemical resistant Viton Bellow material.
• Constant flow of reference liquid,
independent of the process pressure
variations for minimal diffusion potential.
• Suitable for pure water applications and
for polluting fluids.
• Refillable, large KCl reservoir.
Reference
Diaphragm Diaphragm
system
resistance/25ºC
Ag/AgCl (pin)
Sleeve
< 10 kOhm
Silver-silverchloride
Ag/AgCl (pin)
Ceramic
< 10 kOhm
Silver-silverchloride
4.Product & Features
* In application where high process temperature occur together with very low (<2) or very high (>12) pH levels the lifetime is shortened.
Flow at 25ºC
Max. 0.2 ml/day at
10 kPa overpressure
Max. 0.5 ml/day
Table of Contents
TI12B00A20-01E First Edition March 2014
43
w w w.yo ko g a wa.c o m
12) Single pNa electrode:
SR20-AC32
Unit: mm
G 1/4"
G 1/4"
(Yokogawa model: SM23 series)
The SM23 is suitable for long term,
reliable Sodium ion monitoring, another
application is to use the pna sensor as
reference electrode for pH measurement in
combination with and analyzer with dual
high input impedance specification. This
may considerably save maintenance costs in
application where the salt content is constant
and pH>pna+2.
pna measurement is used in many
applications, but the most frequent
applications are found in pure water
applications, where Sodium is measured to
detect carry-over of salts in steam, leakage
of salt through cation filters, Leakage of salt
from leaking condensers.
23
Ø 17
10
Ø 24
K1500FZ
301
Ø 12
128
87
Ø 12
156
SR20DAC32
SR20DAS52
SM23-AN4
26
Wrench 17
6
26
PG13.5
Bellomatic
120
120
4.PRODUCT & FeATURes
Ø 5 20
Ø 24
Bellomatic
a) Additional characteristics for type SM23An4
• Glass resistance @ 25°C 100-300MΩ
• Suitable (Ultra-, pure-) water application
23
Ø17
38
Ø 12
Other applications where the pna electrode
is used for measurement is in brine solutions.
Wrench 17
Please note that accurate measurement
requires that the pH is 2-3 units higher thanPG13.5
the pna value.
Features
• Dimensions and design meet the
requirements of DIn 19263
• Temperature range 0 – 100°C
• Shockproof and heavy-duty (dome
shaped) membranes
• pNa range < 0 to 4
• pH 2 units higher than the pNa
(pH-pna>2)
• na+ range is 0.0001M up to saturation
Unit: mm
SR20-AS52
Ø 12•
Shock-proof glass
membrane with fast
response
15
b) Additional
characteristics for type
SM23-An6
• Glass resistance @
25°C 450 - 700MΩ
• heavy-duty membrane
guarantees great
strength retaining the
necessary sensitivity
• Especially suitable
for the more harsh
processes. For example
brine applications
Unit: mmØ 12 SM23-AN6
SM23-AN4
G 1/4"
Unit: mm
G 1/4"
23
Ø 17
Ø 17
23
15
Ø 12
Ø 12
120
120
Ø 11
Ø 11
Shock proof AN4
Heavy duty AN6
Specifications Single pH Glass Electrodes
Type
SM23-AN4
SM23-AN6
Membrane
in MOhm/25ºC
Universal pNa-glass
bulb (shockproof)
Universal pNa-glass
bulb (heavy duty)
Resistance*
range
100-300
pH
Temperature Reference system Sodium error
range (ºC)
0,1 N[Na ]/25ºC
*
0 - 100
Ag/AgCl (pin)
n.a.
450-700
*
+
0 - 100
Ag/AgCl (pin)
* pNa range < 0 to 4 (0.0001M up to saturation) and pH 2 units higher than the pNa (pH-pNa>2)
Table of Contents
44
TI12B00A20-01E
First Edition
March 2014
n.a.
Back to the pHuture
13) Temperature electrodes:
SM60-T1
Unit: mm
SM60D-T2
Unit: mm
G 1/4"
(Yokogawa model: SM60 series)
For accurate pH measurement temperature
compensation is required. Either a Pt100
or a Pt 1000 temperature electrode can
be selected. However we recommend
to use Pt1000 sensor for more accurate
measurement.
Wrench 17
Ø 17
26
23
PG13.5
Ø 12
Ø 12
120
120
Specifications Temperature Sensor
Type
SM60-T2
SM60-T1
Temperature Pressure
sensor range
Pt100
0 - 1 MPa
Pt1000
0 - 1 MPa
14) Differential pH Sensor:
The sensor responds to pH changes rather
than analyzes the accurate pH value. In that
sense it is best to describe the sensor as
Reference system: pNa, salt sensitive
reference glass. The reference is not a
(liquid) junction but a glass sensor which
responds to the sodium level in the sample
rather than to pH changes (within the
applicable range of the sensor).
Output voltage: The output voltage of
the sensor depends only on the salt
concentration of the process.
Effect of salt: A rule of thumb is that a
change in salt concentration of +/- 25%
has an effect of less than 0.1pH on the pH
reading.
SC24V
Temperature sensor: Pt1000 RTD
(Yokogawa model: PH18 series)
Sensor type: Sodium Reference Differential;
Sterilisable
pH Sensor should be made of pH
sensitive enamel and Reference
sensor be - Na+ sensitive enamel
Construction: Stainless rod with pH sensitive
and sodium sensitive enamel coating
Measuring range be 3 to 11 pH.
Sensor should be able to handle
temperature in the range of 0 to
140 ºC (284 ºF)
Converter requirement: Sensor to be used
in combination with pH meters that feature
Dual high impedance inputs, Adjustable
setting for Isothermal Point and adjustable
temperature coefficient.
18
pH control sensor rather than pH measuring
sensor.
15) Differential pH sensor:
Reference system: No reference fill solution
or junction
Ø6
PH18 series
4.Product & Features
(Yokogawa model: SC24V)
Differential electrode: The differential
measuring principle combines the normal
potential generated by the pH glass with
the potential from a Sodium pNa glass.
In applications where Sodium-, Potassiumor Calcium-salt is present the glass will
generate a stable reference voltage. This
means the measurement can be done
without a conventional reference electrode,
eliminating problems caused by aging and
pollution of the liquid junction.
Temperature
range
-20 - 150ºC
-20 - 150ºC
Sensor should be able to handle
pressure in the range of -1 to 15 Bar
Note: F or accessories like cable
and fittings, pls refer detailed
specification sheets or contact
Yokogawa Respective Node.
Table of Contents
TI12B00A20-01E First Edition March 2014
45
w w w.yo ko g a wa.c o m
5. Lifespan & storage of electrodes
proper performance is guaranteed at first
installation of the product in process if
stored under prescribed conditions. The
proper storage conditions for each type of
sensor are mentioned in the appropriate
instruction manual. On the box the sensor
is delivered in the acceptable storage
temperature and position is indicated by
means of pictograms.
Single reference electrode
When a reference electrode is to remain
unused for a long period it is necessary to
fill the electrode completely with electrolyte
solution, to close the refill opening, and
to protect the diaphragm against drying
out by fitting the protection sleeve. This
protection sleeve must also be filled with
electrolyte solution.
Electrodes of type SR20(D)-AC32 have an
inner duct (see page 43) which contact the
bellow with the process liquid. This duct
must be sealed with the rubber plug.
The best answer to these questions is given
by recommending guidelines for storing
and precautions during usage of the
electrode to maximize the lifetime of the
sensor.
Lifetime is defined as the time the product
shows good performance in the processes.
Because of the variations in process
conditions we cannot be held liable for this
period. However as a guide we can state
that if the lifetime of a particular sensor
has been confirmed in mild conditions the
lifetime in more harsh conditions will limit
the life of the sensor under such conditions.
(See Table in 5.1a)
5.1. Recommended storage
guidelines pH electrodes
5.2. Recommended storage
guidelines
Generally pH sensors consist of a glass
electrode and reference electrode
(junction), of which each of them is a
consumable part. Each has an acceptable
shelf life and life time.
Single pH electrode
When storing a pH sensor, the glass
electrode should be protected with a
cap containing KCl solution to prevent
performance degradation due to drying of
the glass electrode’s glass membrane.
Most customers would like to know what
the expected life time of a sensor is when
used under process conditions. Customers
are also interested in the shelf life for a
sensor when stored in inventory.
5. Lifespan & storage
of electrodes
Since the life of a sensor is dependent
on various parameters like temperature,
pressure, service composition, maintenance
activity taken and sensor handling, it is
difficult to define a particular life. Even for
the same application life may vary from
site to site and will vary from application to
application.
Life time is influenced by many factors
such as sample components, pH value,
temperature, and pressure. Tables 5.1a and
5.1b shows the approximate shelf time and
life time of various electrodes. This is just
a reference for calculating the necessary
consumable amount when purchasing
them. The shelf time and life time are not
guaranteed.
Shelf life and life time
Shelf life is defined as the period in
which a product is saved as stock, and
retains its performance. That means that
The sensors can be stored for
approximately two years, but this is only
a guideline and is not guaranteed. If
the moisture content of the KCl solution
evaporates, causing the glass membrane
to become dry during this storage period,
the performance can be restored by
re-immersing the glass electrode into KCl
solution within the two-year period.
However the proper performance of
the electrode has to be confirmed by
calibration as described in the calibration
chapter applicable to the sensor type.
With the electrodes (types SR20-AS52) the
sleeve must be loosened and lifted. The
hole in the fixed part must be closed by
means of a protective sleeve or tape.
Combined pH sensor
To guarantee optimal performance of
combined electrodes from Yokogawa the
sensors should be stored following the
recommendations given in the applicable
instruction manuals and following the
indications on outside of the original box
the product comes in.
In the table below the shelflife for specific
products are given, however these shelf
life’s cannot be considered as guaranteed
shelf periods.
Table 5.1a
Conditions
Shelf Life
Life Time
Part Name
Normal Temperature/
Pressure
Normal Temperature/
Pressure
80ºC pH4 to 10
Ryton body (PPS
resin)
3 to 5 years
2 to 3 years
1 to 2 years
Glass Electrode
(P/N: K9142TN)
2 years
1 to 2 years
1 month to 1 year
Junction (P/N:
K9142TH)
2 years
1 to 2 years
0.5 to 1 year
KCl Replenish-free
type pH Sensor
(PH8ERP)
KCl Refillable type pH
Sensor (PH8EFP)
Table of Contents
46 TI12B00A20-01E First Edition March 2014
Back to the pHuture
Table 5.1b
Productgroup
Modelcode(s)
Maximum shelf-life in
months
Reference
Reference-electrodes
Flowing type reference (e.g. SR20-AC32)
Non-flowing type reference (e.g. SR20-AC22)
15
Month-code on package
9
Month-code on package
15
Month-code on package
Combined- electrodes
Flowing type combined electrode
(e.g. SC21-AGC55)
Non flowing type combined electrode
(e.g. SC21(D)-AAP26)
Month-code on package
Glass electrodes
SM21*-****(all types)
15
Month-code on package
SM23-****(all types)
15
Month-code on package
SC29*-****(all types)
15
Month-code on package
SM29*-****(alle types)
15
Month-code on package
FU20-****(all types)
9
Month-code on package
FU24-****(all types)
9
Month-code on package
PH20-****(all types)
9
Month-code on package
SC24V-****(all types)
15
Month-code on package
SC25V-****(all Type)
9
Month-code on package
Redox (Combined) electrodes
Four in One electrodes
Notes
5. Lifespan & storage
of electrodes
Table of Contents
TI12B00A20-01E First Edition March 2014
47
w w w.yo ko g a wa.c o m
6. Troubleshooting and diagnostic
6.1 Errors resulting from cracked membranes
A minute crack in the membrane of a glass
electrode is not always visible to the naked
eye. Frequent shocks may give minute
cracks in the membrane with consequently
measuring errors. In this case the analyzer
reading is 0 mV and the analyzer shows a
reading of pH 7, because most pH sensors
using internal buffer with pH 7.
For neutralization processes (setpoint is
typically pH 7) this is very critical and
dangerous situation for environment and
plant are possible. Without additional
diagnostic, the error will not be detected.
Yokogawa is using his patented online
impedeance check. The analyzer checks in
frequently the impedance of pH membrane
(and reference) via the solution ground. In
case of a broken membrane an alarm is
possible to generate.
Note: F or details see chapter “4 Product and
features”
6.2 Fouling of the pH sensor
For industrial applications and particularly for
automatic process measurements it is of the
greatest importance that the sensor is kept
clean.
Fouling of the glass membrane is the
most common cause of error in any pH
measuring system. As the measured potential
is generated on the surface of the pH
sensitive glass membrane it is of the greatest
importance that these membrane surface
remains clean at all times.
When the junction becomes clogged,
contact between the reference system in
the electrode and the process liquid will
be interrupted. This will reduce the speed
of response and may result in unstable
measurement or complete failures.
sensor life time. Afterwards you find example
of automatic cleaning methods.
• Mechanical brushing
A brush actuated by a mechanism
electrically or pneumatically driven, passes
periodically across the membrane.
In most cases it is sufficient to clean only the
electrode before calibration. If it is necessary
to hand clean the electrodes periodically,
suitable cleansers may be used, e.g:
• Diluted solution of hydrochloric acid (10%
HCI solution)
Deposits of hydroxides, lime etc. can be
removed by immersing the electrode in a
diluted solution of hydrochloric acid and
then rinsing with water.
• A solution of water with hydrochloric acid
and pepsin
Aluminous depositions can be removed
by means of a solvent containing: 0,5 ml
concentrated hydrochloric acid, 10 gr
pepsin and 1 litre demineralised water.
After treatment the electrode should be
thoroughly rinsed with clean water.
• Hot water
Depositions of oil and fat can be removed
by means of hot water. If the result is
unsatisfactory, a domestic abrasive may
be carefully used.
But often a periodic frequent cleaning is
needed to prevent building of an layer. In
these cases automatic systems help to reduce
maintenance efforts and help to prolong the
• Jet Cleaner
This jet cleaner gives the opportunity to
avoid build up of deposits and consequent
fouling on the sensor by spraying water (or
air) onto the membrane of the sensor.
• Chamber
Sensor is lifted away from process at
specified intervals and washed in cleaning
chamber. Often possible to keep the
surface of the glass electrode clean by
periodically spraying it with a cleanser.
The cleanser used, its concentration and
the cleaning frequency, depends on the
application. Often it is necessary to sterilize
the electrodes with steam. By means of
steam the temperature of the electrodes is
kept at a certain tempereature (e.g. over
120°C ) for a short time.
•Ultra-sonic cleaning
With the ultra-sonic cleaning method, the
liquid around the electrodes is vibrated.
This is preventing and/or removing any
deposit formed on the electrodes. With this
method optimal cleaning depends on the
vibration frequency and the velocity of the
process liquid past the electrodes.
6.troubleshooting
and diagnostic
Effect of cleaning Contamination by System:
Cleaning system
Process involved
Crystalline scale
Cleaning System
Jet (Water/air)
Brush
Ultrasonic
Chamber
Sugar, fertilizer, soda, glass
B
B
B
A
Suspended matter,
fiber
Ceramic, pulp and paper, metal,
water treatment, wastewater, dairy
B
B
B
B
Viscidity
Flour milling, food processing
B
B
C
A
Algae,
microorganism
River, seawater, industrial
wastewate
A
B
B
A
Absorption deposit
Metal processing/treating,
wastewater treatment (coagulation
sedimentation)
C
AC
A
Rating: A = Good; B = Fair, C = Poor
Note 1: When polar solvents are used for cleaning, these affect the gel-layer
and it is necessary to soak the electrode for some hours to allow
the layer to reform. When a-polar solvents are used first re-clean the
electrode with a polar solvent and after that soak the electrode to
allow the layer to reform.
Note 2: If possible increase the flow rate of the liquid to be measured
past the electrodes. This may result in a reduction of deposit on
the electrodes (typically flow rates of 2m/sec. are necessary).
This method is particularly suitable when measuring pH in liquids
containing solids that settle out quickly.
Table of Contents
48 TI12B00A20-01E First Edition March 2014
Back to the pHuture
6.3 Prevention of the reference
electrode fouling
Clogging or fouling of the reference
electrode can be prevented or reduced by:
a.The correct choice of the junction type
for the application. In highly polluted
liquids the sleeve junction gives better
results than the smaller ceramic capillary
type, alternatively, the reference electrode
with P.T.F.E. capillary may be used to
advantage.
b.Increasing the flow of electrolyte.
By increasing the pressure on the
electrolyte in the reference electrode the
flow will increase. This will reduce the
rate of fouling. The reference electrode
with a flexible viton bellows provides this
feature. For applications where the process
pressure fluctuates widely, and the chance
of process liquid penetrating some types
of reference electrodes is increased, the
“Bellomatic” electrode should be used.
c.Cleaning (See section 6.2 for types of
cleaning)
Clogging of the junction can be remedied
by soaking the reference electrode in 3.3
m electrolyte solution. Slowly increase the
temperature of the solution to max. 80°C
and cool down in the same solution.
This ensures that any electrolyte that may
have crystallized in the junction will become
soluble again.
If the outside of a ceramic junction has been
fouled by the process liquid the junction can
be scoured slightly using fine emery paper,
this increases the flow capacity.
An alterntive cleaning method is to use a
suitable solvent in which the particular solids,
that have fouled in the junction, will dissolve.
The electrode should be rinsed in clean
water before use.
6.4 Poisoning of the reference
electrode
As well as the possibility of fouling, problems
will occur if the reference electrode becomes
poisoned. When the flow of electrolyte in
the reference electrode is insufficient, the
composition of the electrolyte may change
as a result process liquid penetrating the
electrode.
Any change in composition of the electrolyte,
in the electrode reference chamber, may
cause a measuring error.
Some common causes of reference electrode
poisoning are:
a.The reference electrode has been allowed
to run dry allowing the process liquid to
penetrate into the electrode.
This problem mainly occurs with sleeve
junction electrodes, by incorrect treatment,
lack of care in cleaning the ground faces,
or by not re-filling the electrode early
enough.
b.Another frequent cause of poisoning is due
to higher than expected process pressures
or short term high pressure fluctuations.
To aleviate this problem the electrolyte
reservoir should be maintained at a higher
level to increase the electrolyte pressure.
An alternative solution is to replace
the electrode with an integral pressure
compensation type, such as “Bellomatic”
(See page 43/44).
c.When a reference electrode filled with
saturated KCI solution is placed in a
process liquid, which has a temperature
considerably lower than that of the inner
KCI solution of the electrode, considerable
electrolyte crystallization may occur.
d.An excessive formation of crystals in the
electrode may restrict the discharge, and
allow process liquid penetration.
Fig. 6: Salt-bridge (SB20) to prevent
poisoning of the reference electrode.
To overcome this the reference electrode
must be filled with an electrolyte which has
a saturation point higher than the maximum
temperature of the process liquid.
6.troubleshooting
and diagnostic
Note: T he resistance between reference electrode and the process liquid normally
lies between 1 and 10 kΩ at 25°C.
Table of Contents
TI12B00A20-01E First Edition March 2014
49
w w w.yo ko g a wa.c o m
6.5 Errors caused by damped or
electrode cable and connection
It is of the greatest importance to protect
the electrode cables and connections from
ingress of moisture. This may reduce the
insulation resistance of the cable connections
causing a partial short circuit and
consequently an erro-neous reading. Often it
happens that cable is broken.
The online impedance control allows to
detect errors caused by moisture or broken
cable.
Note: In case of moisture problems Yokogawa
can offer easily solutions to solve this
item, like SENCOM (SENsors with
digital COMmunication)
6.6 Errors caused by shorting the
electrode cables
Often, it is usual for installation technicians to cut
off surplus cable lengths. This is done despite
the availability of different cable lengths.
The shortening is strongly to be dissuaded.
To prevent “inside radiation” of disturbances
the cables are supplied with a special layer
of graphite for screening. This layer can be
removed very difficult but after shortening
the cable length regularly measuring errors
are stated (e.g. instability) or are strongly
“manual” sensitive. The cause of the
measuring error was the result of a strongly
decreased isolation resistance between core
and screen of the electrode cable.
For cables of the glass electrode this
isolation resistance must be over 1000 times
the resistance across the glass membrane.
6.7 Interferences by stray voltages
in the liquid
In many industrial applications the voltage
potential of process liquids is significant
and cannot be neglected when making pH
measurements. This voltage level may be
due to a number of causes, a full study of
which is too complex to be covered in this
book. It is essential therefore, that the liquids
are earthed at the point where the pH value
is to be measured.
This can be easily achieved when metal
fittings are used. When plastic fittings are
used they should be fitted with an earthling
electrode of suitable metal.
6.8 Errors caused by poor
installation
When commissioning a pH measuring
system, care should be taken that the
electrode connectors do not become wet,
or fouled with fine deposits. Particular
care should be taken if the electrodes are
removed for tests, or pressure testing of the
pipe, and are left unprotected.
If this occurs any dirt should be carefully
removed from the electrode connections and
the cables removed and, either thoroughly
dried, or preferably replaced.
6.troubleshooting
and diagnostic
Notes
Table of Contents
50 TI12B00A20-01E First Edition March 2014
Back to the pHuture
I. ReLATeD TO sensORs:
7.1. What do we do when the shelf
life of pH sensor expires?
In general conductivity is a better and more
accurate method to measure the salt concentration.
The shelf life of the sensors is defined as the
period that the sensor can be stored without
loss of functionality. It does not mean that
the sensor is not functional after this time.
Inspection before installation is highly recommended. If the wet pocket still contains electrolyte then buffer check prior to installation
is needed. If the response in the buffers is
slow then regeneration of the sensor is needed: follow instruction manual.
If the wet pocket is dry, then there is a good
possibility that the internal reference electrolyte has evaporated and the sensor is not
functional anymore.
7.4. What is an ion selective electrode?
7.2. What is the best sensor to measure pH with in a 0 to 20% HCl?
Typical maximum flow rate is 5 m/s.
However it depends on the application. In
pure water applications flow rates must be
as low as possible, because the sample
flow will wash out the electrolyte from the
junction. When the sensor is mounted perpendicular to the process flow the sensor
will experience vortex vibrations and these
can seriously shorten the life time.
7.3. How can I measure salt concentration with a pNa electrode?
We use the pna sensor normally as a
reference electrode for pH measurement.
It is also possible to use this electrode as
a measuring electrode in concentrated salt
solutions.
A general rule is that a pna sensor only
responds to pna, when the pH value is 2
units higher.
For example the requested range is 1- 100
g/l naCl. It is a logarithmic measurement
so the range really is 1-100 g/l. The molecular weight of naCl is 60. The logarithm
of the minimum naCl concentration is log
1/60 = -1.78. So the pna value is 1.78.
It means that this range we can measure
with the pna sensor when the pH is
higher than 4. If the requested range
is 0- 100 mg/l naCl then the same
calculation shows that the measurement can
be done only if the pH is higher than 7.
7.5. What is the best sensor to pH
with in a 0 to 20% NaOH?
Same answer as for question 7.2 basically.
On top of this the pH sensor suffers from
high alkaline errors at values above pH 13
(4 g/l naOH).
7.6. What is maximum flow rate for
a pH sensor?
7.7. Why do you sometime find pH
range specifications typically 2- 12?
every pH sensor measures under reference
conditions (atmospheric pressure, ambient
temperature) accurate over the 0- 14 pH
range. However in real life with high temperatures and fluctuation process pressures
the lifetime of the sensor can be considerably lower especially with gel filled reference cells. Therefore we normally limit the pH
range for gel filled sensors to 2- 12 pH.
The electrolyte must flow to the process with
a flow rate that prevents ingress of process.
Generally an overpressure of 0.1 barg =
1 m water column is sufficient, but it must
be guaranteed at all times. Only in case of
Bellomatic you have a constant overpressure
of 0.3 barg independent on the process
pressure.
In all other cases the pressure on the
electrolyte must be higher than the maximum
process pressure: static and dynamic
pressure. The static pressure can be
estimated easily, but the dynamic pressure is
difficult to predict. It depends on viscosity and
flow rate of the process and the dimensions
of the downstream process piping.
7.9. How can I create a steady
flowing reference for processes with
varying pressure?
Only Bellomatic sensor offers this function.
An alternative is a regulated pressure on the
electrolyte reservoirs which is a costly solution.
7.10. What are the advantages of
3.3 molar KCl ?
All pH sensor suppliers use a saturated or
a concentrated KCl solution in the reference
cell. The advantage is that a concentrated
solution decreases the electrical resistance of
the reference cell.
The solubility of KCl in water is 238 g/l at
0ºC, which is approximately 3.3 mol/l.
Therefore the advantage of 3.3mol is that
the solution stays below saturation when the
temperature drops. The crystals would stop
the flow and block the junction.
Eh = Half Potential (Ag/AgCl in KCL) against SHE
290
270
250
mV
The best method to measure a range of 020% HCl is Inductive Conductivity. There we
can reach an accuracy of 2% of the conductivity. A pH sensor is not a very good
solution. The strong acid solution tends to
generate high diffusion potentials and even
if the pH measurement is 0.1 pH accurate
(logarithmic), then the error in concentration
(linear) is 25% relative
See question 7.3. ISe is never selective: It
just responds to changes in the ionic concentration. This is expressed in the specifications
of the ISe as selectivity coefficient.
7.8. When I am using a flowing sensor,
what is the overhead pressure needed?
7.FReQUenTLy AskeD
Q&A
7. frequently asked Questions & answers
230
210
190
170
150
fig.: 7.11
0
1
2
Molar
3
4
5
Table of Contents
TI12B00A20-01E
First Edition
March 2014
51
7.Frequently Asked
Q&A
w w w.yo ko g a wa.c o m
7.11. What happens if we combine a
measuring electrode with 3.3 mol KCl
electrolyte with a reference electrode with
1 mol KCl in a pH sensor or vice versa?
In the figure 7.11, you see that the reference
potential difference between 1 and 3.3
molar is approximately 30 mV.
The sensitivity of a pH sensor at reference
temperature is 59mV/pH. So if you combine
a measuring cell filled with 3.3mol with
a reference cell with 1mol KCl, then you
will have an offset (asymmetry potential)
30/59mV = 0.5 pH approximately.
This unbalance will be solved when the
sensor is calibrated and it will show itself as
Asymmetry Potential.
7.12. What is the purpose of high
viscosity KCL solution?
The viscosity of the electrolyte decreases
with increasing temperature. When you
have a Bellomatic that lasts 2 months
before it needs to be refilled at reference
temperature and you are measuring at
o
80 C, then the viscosity is only 40% of
the original value. The sensor needs to
be refilled every 3 weeks. This is too
much maintenance and in that case we
recommend replacing the electrolyte by the
electrolyte with increased viscosity.
7.13. What is the typical life time of
a buffer solution and after the bottle
is open?
When the bottle is open, the electrolyte is
exposed to air. Air contains 0.04% CO2
and this gas is absorbed by the buffer solution as Carbonic acid. If you expose pure
water to air, the pH will drop to 5.5 pH. It
means that buffer 4 and buffer 6.85 are not
very sensitive. The buffer quality does not
degrade strongly when the bottle is opened.
The alkaline buffers are much more sensitive
to exposure to air and the pH will continue
to drop. So the buffer 9.18 must be kept in
a closed bottle and cannot be kept long
after opening.
7.14. How many pH analysers can I
calibrate with the same buffer solution?
We recommend using pH buffers only
once. If you want to use the same buffers
for more analysers, then you must avoid
contamination of the buffers by the sensors.
The sensor must be cleaned and washed
carefully before inserted into the buffer.
A sensor is often cleaned with acid or with
detergent to remove the fouling or the scaling. After this chemical cleaning it must be
washed with demineralized water. This must
be done three times with fresh water before
it can be considered clean.
This must also be done between the two
buffer solutions.
Then you can calibrate up to 10 sensors with
one set of buffers as long as it is in a short time.
The difference with Pt100 is that the value
varies only 0.391 Ω per ºC.
If we connect the Pt100 with a pH cable of
10 m then the resistance of this cable will
be 0.7 to 0.8 Ω which results in a +2 ºC
measuring error on the temperature signal.
If we connect a Pt1000 sensor, then
the sensitivity is 3.91 Ω/ºC. So when
connected to the same cable the measuring
error is only +0.2 ºC. Good enough for pH
measurement.
7.16. Can we use a glass electrode
as reference? If yes, how and
where?
When you use Glass electrode as reference,
we call the pH sensor a differential sensor.
There are three different examples:
7.15. What is the difference between
Pt100 and Pt1000 temperature sensor and which one is better?
Both sensors have the same temperature
function and follow the following table; The
Pt1000 values are simply a factor 10 higher.
see table under: Resistance value of Pt 100
sensor at various temperatures.
a:pH compensated ORP measurement:
This is the most common example.
ORP depends on the pH value as the
NERNST-Clark equation tells us: E=
RT/2F * ln [H+]/[H2]. –log [H+] is the
pH and in the same way we define –
log[H2] as rH. Then rH= 2*Eh/59.16 +
2.pH. It means that solutions with equal
oxidizing power (rH), the ORP varies 59
mV at each pH unit change. So does
the pH measurement electrode. So, if
we use a pH electrode as reference we
achieve automatic compensation of the
ORP measurement.
b:Differential pH measurement: Some
manufacturers use a pH glass electrode
in a buffer solution and use this as
a reference cell. The output of the
reference pH cell is independent on the
ºC
0
1
2
3
4
5
6
7
8
9
0
100.000
100.391
100.781
101.172
101.562
101.953
102.343
102.733
103.123
103.513
10
103.902
104.681
104.681
105.071
105.460
105.849
106.238
106.627
107.016
107.404
20
107.793
108.181
108.570
108.958
109.346
109.734
110.122
110.509
110.897
111.284
30
111.672
112.059
112.446
112.833
113.220
113.607
113.994
114.380
114.767
115.153
40
115.539
115.925
116.311
116.697
117.083
117.469
117.854
118.240
118.625
119.01
50
119.395
119.780
120.165
120.550
120.934
121.319
121.703
122.087
122.471
122.855
60
123.239
123.623
124.007
124.390
124.774
125.157
125.540
125.923
126.306
126.689
70
127.072
127.454
127.837
128.219
128.602
128.984
129.366
129.748
130.130
130.511
80
130.893
131.274
131.656
132.037
132.418
132.799
133.180
133.561
133.941
134.322
90
134.702
135.083
135.463
135.843
136.223
136.603
136.982
137.362
137.741
138.121
100
138.500
138.879
139.258
139.637
140.016
140.395
140.773
141.152
141.530
141.908
Table of Contents
52 TI12B00A20-01E First Edition March 2014
Resistance in ohm
Resistance value of Pt 100 sensor at various temperatures.
Back to the pHuture
7.17. What is smart pH sensor and
what are the advantages?
A smart pH sensor has an amplifier in the
sensor and translates the information into a
digital signal that is transmitted to the pH
analyzer. With the SENCOM we measure
the mV readings of the three elements of
the pH sensor plus the resistance of the
temperature element. We also measure the
impedances between these elements and
calculate pH, ORP, impedance values. The
smart sensor does everything a pH amplifier
does and the information is transmitted
through a serial communication. The sensor
stores latest calibration data and uploads
these data to the pH analyzer.
Advantages are:
a:The sensor can be calibrated in the
laboratory and stored. As soon as it
is connected to the pH analyser in
the field, these calibration constants
are uploaded automatically. No field
calibration is necessary.
b:The cabling between sensor and
analyser is low impedance cabling, so
insensitive for noise and humidity.
c:Due to the fact that digital signals are
transmitted, it becomes possible to
communicate over a longer distance (up
to 60 meters).
7.18. Is it possible to connect competitive
sensors with Yokogawa analyzers?
If yes, what are the conditions?
All Yokogawa analysers are using an open
architecture with adjustable temperature
compensation, adjustable isopotential point,
adjustable slope; they are compatible with
all direct pH or ORP sensors. The only
exception is SMART sensors. If the sensor has
a built-in amplifier, it cannot be connected.
7.19. Can I connect pH sensors without
solution ground? If yes, how?
Yes we can. but we do not recommend
this method. A pH sensor has typically a
reference cell that is low impedance to the
process. If the process suffers from common
mode voltages then these voltages tend
to generate ground loop currents. These
currents flow through the path of least
resistance. which is the reference cell. This
results in measuring errors (Ohm’s law: 1
uA through 10 kΩ is 10 mV is 0.2 pH) and
damage of the reference cell: a battery
connected to an external power source is
not good!
With solution grounding the path of least
resistance is the solution ground and the
pH sensor does not suffer from ground loop
currents
Also the famous Yokogawa impedance
monitoring features work properly only when
we have a solution ground in the sensor.
If you want to connect a sensor without
solution ground, then you shorten out the
reference cell amplifier by jumpering terminal
13 (input 2) with 14 (common voltage of the
analyser).
and is recommended for continuous use in
high temperature processes. Under these
conditions the lifetime will be longer and the
stability better resulting in less maintenance.
The “price” of this superior specification is
the electrical impedance. This impedance
doubles with every decrease of temperature
of 10 degrees. so when you calibrate such
a sensor in the wintertime the response
is rather slow. The impedance on the
specifications is the impedance at 25
degrees Celsius and it is not wise to use
sensors at temperatures where this value
exceeds 1 GΩ= 1000 MΩ.
7.22. What is the benefit of using
heavy duty glass?
Heavy Duty Glass is the description of pH
sensors that feature a wall thickness of the
sensitive glass membrane of approximately
1 mm. So all customers that consider pH
sensors as fragile will benefit from this
feature. Also applications corrosive to Glass
like hot alkaline solutions with high salt
content or processes with risk of HF (Fluoride
containing waters at low pH) will benefit
from heavy duty features.
7.20. When should we use salt
bridge over flowing reference?
7.23. What sensor should I choose
for wastewater that contains oil?
Flowing reference cell and salt bridge have
the same purpose: to prevent diffusion
through the junction and fouling of the sensor.
Sometimes a salt bridge is easy to retrofit
existing installation where you experience
sensor troubles. You keep the same
reference sensor but place it in a salt
bridge.
pH sensors are not suitable for measuring
pH in oily water. because the oil will
generate a coating on the pH membrane
and make this membrane insensitive.
Also the oil will penetrate the liquid junction.
This means that sample preparation is
necessary when measuring oily water.
If the process reacts with KCL or cannot
be contaminated with KCl you can use a
salt bridge filled with another solution like
NaNO3.
In high temperature applications the
maintenance may be even less with a salt
bridge. because the reference cell is under
reference conditions. Refilling of the large
reservoir is easy and does not need to be
done frequently.
7.21. When do I choose G type
glass over L type?
G-Glass is the best general purpose
pH glass. It couples a good chemical
resistance with a wide measuring range
and relatively low electrical impedance.
L-Glass has a better chemical resistance
7.Frequently Asked
Q&A
pH. Because it is inserted in a buffer.
The buffer is in contact with the process
through a junction. The advantage
is that the sensor is hermetically
sealed and cannot be poisoned.
The shortcoming is that the pH buffer
develops diffusion potentials in the
junction.
c:Salt based reference: This principle we
use with SC24 sensors. The reference
cell responds to changes in salt
concentrations and if it does not respond
to changes in pH value then it is the
perfect sensor.
The oil is typically lighter than water so a
phase separation based on gravity is the
best solution. In open channels it must be
avoided that the sensor measures at the
surface and regular cleaning coupled with
the dynamic response diagnostic function is
recommended.
7.24. Can I improve the life of pH
sensor when I choose 225 mm
electrode in place of 120 mm
electrode?
That is well possible. because the 225 mm
sensor has more electrolytes. So depletion
by diffusion will take a longer time. On
the other hand it is more likely to break by
mechanical load like vibration or during (dis)
assembly.
Table of Contents
TI12B00A20-01E First Edition March 2014
53
7.Frequently Asked
Q&A
w w w.yo ko g a wa.c o m
7.25. Can we mount pH sensors
upside down?
Most of the pH sensors with glass membran
cannot be mounted upside down. To absorb
the thermal expansion of the internal buffer
there is always a considerable air bubble in
the sensor.
II.
When the sensor is mounted upside down
the reference element can lose the contact
with the electrolyte. because it inserts in air
instead of electrolyte.
The FU24 pH sensor from Yokogawa
can be used in upside down application,
because it have a special design with
reduce the size of the air bubble.
7.26. Does Yokogawa offer a nonglass pH electrode? If not why?
Yokogawa offers the PH3A and PH18
sensor. These two type of sensors are nonglass and suitable for regular CIP and SIP
cycles.
RELATED TO Analyzers/ CONVERTERS:
7.27. Why does the FLEXA has 7 different
options for temperature sensor configurations?
One of the features of EXA and FLEXA
is the open architecture: One unit for all
applications; one unit for all sensors. The
most common difference between the
various pH sensors is the temperature
compensator. We want to be compatible
with most direct pH/ORP sensors on the
global market.
The following temperature compensators are
supported by EXA/FLEXA.
Pt1000 is used by Yokogawa in recent years
Pt100 is used by many European
manufacturers like Yokogawa, E&H, MT and
Hamilton.
5k1 is used by Yokogawa Japan in the past
3k Balco is used by US suppliers like ABBTBI and Rosemount
8k55 is used by Honeywell-Leeds & Northrup
350 and 6k8 were used by Yokogawa and
PTC 10k were used by US suppliers like
Inventis-Foxboro.
7.28. Why is backlit not available in
the 2 wire analyzers?
The 2-wire analyzers have a range of 4- 20
mA at a supply voltage of 16 V. That means
that we have only 4*16= 64 mW power
available. With HART communication and
BURN-OUT feature we have only 50 mW.
These 50 mW must power the preamplifier
and the CPU plus the display.
This leaves little room for extra’s like backlit.
7.30. Can we select jet cleaning
along with 2 wire analyzer?
7.29. What is a 2-wire analyzers &
what is a 4-wire analyzers?
Which one suits your application?
It is possible to use PH201 for chemical
washing of the sensor. The PH201 uses
a digital signal on the 2-wire line, so in
that case HART is not available. PH201 is
General Purpose only and does not have
CE and CSA approval.
This terminology only refers to power supply
of the analyzer. It does not have any
reference to the number of wires.
The 4-wire analysers separate the power
supply from the measurement output. As a
minimum we need two wires for the power
supply and two wires for the current output
of the analyser. The PH450 has two analog
outputs. four digital outputs. One digital
input and of course the sensor input wiring.
That is why PH450 has 6 cable glands.
The advantage is that any functions are
combined in one enclosure. Local control
and alarm functions are possible. In small
pH control installations this is a cost saving
feature.
This combination is possible with PH201
distributor. The jet cleaning requires a signal
HOLD function and a start wash signal. The
PH201 uses a special digital signal on the
mA wiring, so it cannot be combined with
HART communication.
7.31. In a dual input FLXA21
analyzer, how can I extract output
from the second input module?
In a dual input FLXA21 analyzer you can
use a P&F HART® to convert the HART®
dynamic variables into current outputs or
contact outputs. The model name is KFD2HLC-Ex1.D
The 2-wire analysers commonly referred
to as Analyzers combine the wires for
current output, for power supply and digital
communication.
The FLXA21 combines the power supply
with mA output and HART® communication:
all with only two wires plus shield. The
advantage is safety (only 24VDC) and
simple installation. Just two wires between
control room and analyser. This is an
advantage in large scale operations where
process control is centralized in the DCS
system.
Table of Contents
54 TI12B00A20-01E First Edition March 2014
Back to the pHuture
III.
7.32. Can we use a pH analyzer as a
temperature analyzer? If so, how?
The pH sensor has an accurate temperature
sensor for temperature compensation. With
PH450 it is easy to select one mA output for
temperature. Also with FLXA21 it is possible
to read temperature as dynamic variable on
the HART converter see Q. No. 7.31.
This information is useful to understand the
pH measurement and in case of a bypass
or extractive measurement it is a good “flow
checker” because good flowing sample has
higher temperature than when flow rate is
too low (when the sample flow is blocked.)
So it is a diagnostic tool.
7.Frequently Asked
Q&A
In FLXA21 you can select which dynamic
variables you like as SV. TV and QV:
Choice for FLXA21-pH is Temp-1; Ref imp-1;
pH-2; temp-2; ref imp-2; calculated value;
redundant as example of the flexibility of
FLXA21.
Commissioning>Advanced Setup>HART>
7.33. Why do we choose differential
amplifier? See Q.7.19.
COMMON (related to both sensor and analyzers) QUERIES AND ANSWERS:
7.34. What feature does an
analyzer need to connect an ion
selective electrode?
The two important features/functions are:
Possibility to change the Isopotential point
and Possibility to generate a concentration
table to have output linear to concentration
Assume the range of the analyser is requested
to be 1- 100 g/l NaCl. and then the output
range is -0.22 to 1.78 pNa/pH. See Q3.
Now we use the concentration mode for salt
concentration and use a 21 point table function.
Next you define the concentration scale. Here
the range is reversed. because the higher the
pNa value is the lower the concentration is and
you want to have a range of 0- 100 g/l which
is 0- 100 ppt (parts per thousand= g/kg)
7.35. Can I connect ion selective
electrode to Flexa?
The name Ion Selective Electrode (ISE) is not
correct. because every ISE is also sensitive
to other ions than the one mentioned on the
specification sheet. Take for example the
pNa electrode. This responds to Sodium ion
concentration if the pH is higher than pNa +2
and if no other salt ions are present. So the
name Ion Sensitive Electrode is better.
Assume that this is all OK. and then we
connect the ISE to the input 1 terminals of the
FLEXA and a reference electrode to the input 2
terminals. When we use as example the pNa
sensor (SM23-AN4), then we know that the ITP
is 0. (The inner buffer solution is 1 mol/l NaCl)
so we program ITP as 0 and calibrate the
analyser with a 1 mol solution (pNa= 0) and
a 0.01 mol solution (pNa= 2). The display in
the FLEXA displays only knows pH. so we read
pNa for pH.
IV.
fig. 7.35: linerarisation function of analyzer
RELATED TO MEASUREMENTS:
7.36.Why does the process
measurement and laboratory
measurement not match?
It is not so, that they do not match. You
just do not hear it when they do match.
Then everyone is happy. When they do
not match, there is a problem and it is
natural that the in-line analyser is considered
the “wrong” one, because the laboratory
is the reference method and the in-line
measurement is just a tool to control the
process.
Anyway when they do not match, you must
find the reason:
1. A
re both analysers accurate? To check
this both analyzers must be validated,
not calibrated. So you take 2 or 3
fresh buffer solutions and measure these
solutions with both analyzers. Do not
make any adjustment. Please take a
piece of paper and write down the
values and judge the results. If one of
them is wrong by 0.1 pH or more then
that one needs to be calibrated. Once
you have done so, you must repeat the
validation test.
2. C
ompare apples with apples: The
in-line measurement and the laboratory
analyser must measure the same sample
at the same pressure and the same
temperature. In other words, bring the
teapot to the kettle and not the kettle
to the teapot. It is possible to measure
at the lab at reference temperature
and the in-line measurement at process
temperature. But this can only be done
when the in-line analyser is properly
compensated for temperature. You
can check this by taking a hot sample,
Table of Contents
TI12B00A20-01E First Edition March 2014
55
7.Frequently Asked
Q&A
w w w.yo ko g a wa.c o m
insert the sensor in hot sample and let
the sample cool down to 25 degrees.
Only when the reading does not
change, you have proper temperature
compensation. If the reading changes,
calculate the change in delta pH /
delta temperature and program this
coefficient in the (FL)EXA
3. H
ave reasonable expectations. If both
analysers have an accuracy of 0.1pH,
you can accept differences in readings
up to 0.2 pH.
4. T ake into account properties of the
process. As example we take boiler
feed water. This sample is ultrapure
water with traces of Ammonia or
Morpholine to increase the pH. This
sample is completely unbuffered. So
as soon as it is exposed to ambient air
the pH will drop due to absorption of
Carbon dioxide from the air.
V.
7.37. What do you recommend to
use as a laboratory instrument?
The PH72 is an excellent device for
validation of the process analyser. It is
small. accurate and it allows you to do the
validation measurement as close as possible
to be able to compare apples with apples.
7.38. How accurate is your pH
analyser?”
Generally we state that the pH analyser
is as accurate as you are. With lots of
care you can achieve 0.05 pH accuracy.
With normal care you achieve 0.1 pH and
without care the measuring error can be pH
0.5 or more. More information you find in
Q7.36 and Q7.37
This is difficult and can only be done
through additional device like Field mate
through HART communication.
7.40. Can pH and ORP be measured
simultaneously?
Most of our pH sensors have a noble material
solution ground and the voltage of this is
measured independently from the voltages of
the pH cell and the reference cell.
So we can also utilize this voltage to get
you ORP readings. The beauty of this
solution is that you can choose with one
sensor and one (FL) EXA for pH analysis. pH
and ORP analysis and just ORP analysis.
Note: B
uffered means that weak acids or weak
alkaline are in the water that absorb ingress
of acid or alkaline contaminants. Examples
are Carbonates and Phosphates. When such
a solution is contaminated with an acid then
the pH does not change much. This is principle of pH buffers. Un-buffered means that
no weak acids are in the water, so ingress of
acid will immediately change the pH value.
7.39. Is it possible to record process
and diagnostic data on a personal
computer? If so what are the methods?
MAINTENANCE. SERVICE & REPAIR RELATED:
7.41. When do we recommend
automatic cleaning / calibration?
Good pH sensors only need monthly
validation. so no need to do automatic
calibration as long as the customer has
access to the sensor. In some applications
the sensor gets fouled seriously during
usage and then automatic cleaning devices
like chemical spray unit or auto retractable
fittings can be installed successfully.
Typical customer experience with fully
automatic auto calibration systems is that the
system needs more maintenance than the
pH sensor used to need. Plus the autocal
procedure often results in mis calibration of
the sensors. (Mixing buffers. poor washing
of calibration chamber. empty buffers
bottles)
7.42. When someone states that pH
sensor is failed, what details are
needed for evaluation?
When a sensor has failed, then following
information are needed for evaluation:
model code, serial number, application
details, description of the failure, lifetime
stand time and diagnostic data from
analyzer (slope, asymmetric potential, glass
impedance, reference impedance).
Then if it is clear from the details that
sensor needs to be sent to the factory for
investigation, please ask for the document
called “Clean Sensor Declaration” to protect
the health and safety of our colleagues
in the laboratory. Additional relevant
information other than the queries asked
is also helpful. For example; what are the
validation results? What does the sensor
read in 3 different fresh buffer solutions?
7.43. How can be the process
reading be wrong when it is correct
in the pH buffers?
This is the well-known pH problem that we
call Diffusion Potential. If the sensor junction
is plugged, then there is no good electrical
contact between electrolyte and process.
This results in diffusion potential that is
directly measured as error. The chemical
composition of pH buffers is different from
process liquids. So when the junction is in
bad condition, you calibrate for this error in
the pH buffers, but they are different from
the process. An easy check is to look at the
diagnostic information on the pH analyser:
Is the Asymmetry Potential high or the Slope
low. then most likely you have this problem.
Another cause can be the infamous ground
loop current. This can be the case when you
use pH sensors without solution grounding
7.44. Can a reference sensor be
stored in demi-water?
Sensors that are not used need to be stored
in a solution that guarantees that the sensor
is ready for use.
When the sensor is stored in demi water the
salt will be washed out of the junction of the
reference cell. So it is not a good solution.
Yokogawa uses the same salt concentration
in the wet pocket as is used inside the
sensor.
For combination sensors we add a trace of
acid to keep the Glass membranes active.
So best method is to keep the wet pocket
and store the sensor back in the wet pocket
when not in use.
7.45. Is there any special
consideration for pH sensor storage
& transportation?
The standard packaging is designed for
long time storage. So it should be used
when long storage is foreseen.
We experience most troubles with storage
in wintertime when service engineers keep
sensors in the trunk of the car while the
outside temperatures are below the minimum
allowable temperature of -10 degrees. Also
we experience sensor failures if the sensors
are shipped through air with planes that do
not have conditioned storage compartments.
Table of Contents
56 TI12B00A20-01E First Edition March 2014
Back to the pHuture
7.49. What’s the best calibration set?
The best set is obviously the NIST set: 4.01;
6.87; 9.18. They are standards all over
the world and formulated by Mr. Bates of
the NBS many years ago. They couple a
high level of buffer capacity. dilution value.
and minimal junction potential to wide
availability.
A general rule is that analysers need to be
validated once per month. Only when the
validation shows that the accuracy is out of
its required limits, calibration with buffers is
needed after cleaning of course.
After this calibration the sensors need to be
validated of course.
All other sets are convenient for some users
but do not have the same quality and
standardization.
The couple 4 and 6.87 is most stable and
should be used for calibration. The other
values can be used for validation purposes.
7.47. Do I need to re-calibrate after
cleaning my pH-electrode?
This is a widely spread misconception.
When a sensor needs cleaning it needs to
be cleaned and only when a sensor does
not meet the validation requirements after
cleaning it needs to be calibrated.
7.48. How do I calibrate to get
accurate pH and accurate ORP
reading in the same sensor?
All (FL) EXA analysers have independent
calibration routines for pH and for ORP.
So at first you calibrate the pH sensor
function using the autocal function with
buffer 4.01 and pH 6.87. Then you use
manual calibration for the ORP function.
The ORP calibration is normally a one point
calibration. although a 2 point calibration is
possible.
7.Frequently Asked
Q&A
7.46. How often do I need to
calibrate my pH system?
Proceed as normal then repeat:
For calibration you would use normally a
chinhydron solution of which the ORP value
depends on the type of reference cell.
Please understand that when you have a
pH sensor as reference cell the value would
always be the 7.00 pH value, because
this is the reference pH value (88 mV in
saturated Ag/AgCl like is used in FU20).
Commercially available ORP buffer solutions
do not specify the pH value. so they cannot
be used to calibrate ORP analysers when
pH sensor is used as reference cell.
Note: M
ore detailed information you found in
chapter 3.8.
Table of Contents
TI12B00A20-01E First Edition March 2014
57
7.FReQUenTLy AskeD
Q&A
w w w.yo ko g a wa.c o m
REMOVE
ELECTRODE CAP
YOKOGAWA
Electrolyte
BELLOW DOWN
Fig. 7.50: Refill Bellomatic (SR20(D)-AC32)
7.50. How to refill a bellomatic
sensor?
There are several methods described in the
manual and other publications.
There is also the easy method that requires
only a pencil. We assume the customer
kept the original package. In there you find
the rubber plug that sealed the inner tube
prior to installation:
Step 1: remove the electrode cap: same as
first drawing in fig. 7.50.
Step 2: Push the bellow down with the
backside of the pencil and insert the rubber
plug in the inner tube at the bottom of the
Bellomatic.
Step 4: Fill the sensor with the proper
electrolyte leaving 3 cm from the top without
electrolyte.
Step 5: Assemble the electrode cap.
Step 3: Remove the pencil. The bellow will
stay down if the plug is in place.
How to refill solution in bellomatic sensor?
For more details, see the instructions on YouTube:
https://www.youtube.com/watch?v=M9-nUhWBGUM)
Table of Contents
58
TI12B00A20-01E
First Edition
March 2014
Back to the pHuture
7.Frequently Asked
Q&A
Notes
Table of Contents
TI12B00A20-01E First Edition March 2014
59
w w w.yo ko g a wa.c o m
Appendix 1: Chemical Compatibility
pH Compatibility Chart
Sulfuric Acid
Inorganic Acid
Hydrochloric Acid
Organic Oxidizing
Neutral Salt
Solvent Agent
Basic
Salt
Acid Salt
Alkali
Organic Acid
Appendix1: Chemical
Compatibility
It is the customer’s responsibility to decide
the material. First of all, please confirm what
material customer use in their sample.
O=
*=
X=
Nitric Acid
Phosphoric Acid
Boric Acid
Carbonic Acid
Chromic Acid
Sulfurous Acid
Acetic Acid
Formic Acid
Oxalic Acid
Lactic Acid
Phenol
Monochloroacetic Acid
Calcium Hydroxide
Potassium Hydroxide
Sodium Hydroxide
Ammonium Hydroxide
Ammonium Chloride
Zinc Chloride
Iron (III) Chloride
Iron (III Nitrate)
Sodium Sulfite
Sodium Carbonate
Sodium Phosphate
Potassium Chloride
Sodium Sulfate
Calcium Chloride
Sodium Nitrate
Aluminium Chloride
Hydrogen Peroxide
Sodium Hypochlorite
Chlorinated Lime
Potassium Dichromate
Alcohol
Organic solvent or oil
(excluding alcohol)
Chlorinated Solvent
Can be Used
Shortens Useful Life
Cannot be Used
These tables should be used only as a
guide and no guarantee is given. The tables
show the chemical resistance of material to
individual chemicals. The result may differ if
a sample contains more than one chemical.
It is recommended to refer to multiple tables.
Concentration & pH
Material
W/V (%) pH (25ºC)
0.5
1.0
0.05
2.0
0.4
1.0
0.04
2.0
0.6
1.0
0.06
2.0
1.0
1.5
0.6
5.0
0.6
3.6
1.2
0.8
0.8
1.4
0.6
2.8
0.5
2.3
0.9
1.0
0.9
2.4
0.9
5.4
0.9
1.8
0.2
12.4
0.5
12.7
0.4
12.9
0.5
10.4
5
5
5
5
1.3
5
5
11.8
5
5
5
5
5
8.2
5
1
1
12.5
1
5
4.5
10
PVC PVDF PP PTFA 316 SS Ryton Silicone Viton EPDM
O
O
O
O
X
O
O
O
O
O
O
O
O
X
O
O
O
O
O
O
O
O
X
O
O
O
O
O
O
O
O
X
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
*
O
O
O
O
O
O
O
O
O
O
O
O
*
O
O
O
*
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
*
O
O
O
O
O
O
O
O
*
*
O
O
O
*
*
O
O
*
O
O
*
O
*
*
O
O
O
O
O
*
O
*
*
*
O
*
*
O
O
O
*
X
O
O
O
O
O
O
O
*
O
*
O
O
O
O
*
O
X
O
O
O
O
*
O
*
O
O
O
O
O
O
O
O
O
O
O
O
O
O
*
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
X
O
O
O
O
O
O
O
O
X
O
O
O
O
O
O
O
O
X
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
*
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
*
O
O
O
O
O
O
O
O
*
O
O
O
O
O
O
O
O
*
O
O
O
O
O
O
O
O
*
O
O
O
O
O
O
O
O
*
O
O
O
O
O
O
O
O
X
O
O
O
O
O
O
O
O
*
*
*
O
O
O
O
*
O
X
*
*
O
*
*
O
O
O
*
*
*
O
*
O
O
O
O
*
O
O
O
O
O
O
O
O
O
O
O
*
O
*
X
O
O
*
X
O
O
O
*
O
*
*
*
Notes:
The recommendations of this chart are based on evaluations of
"single components" processes. You should consult your plant
Table of Contents
60 TI12B00A20-01E First Edition March 2014
O
O
O
*
Chemical Resistance Table
Back to the pHuture
This table should be used only as a guide and no guarantee is given. The table shows
the chemical resistance of material to individual chemicals. The result may differ if a
Table 9-2 (from
TI 12B07A03-01E)
sample
contains more than one chemical.
: Excellent
: Good
: Not so good
X : Unusable
Sulfurous acid
Inorganic acid
100 20
90
20
5
80
5
6
30
5
30
10
b
14
30
Hydrobromic acid
Sulfuric acid
Phosphoric acid
Ammonia water
20
40
20
40
10
10
57
57
3
3
30
30
15
15
20
80
20
70
20
100
60
100
80
100
20
20
80
100
100
X
Alkali
Potassium hydroxide
Sodium
hydroxide
Sodium hydroxide,
Sodium hydroxide9 to 11%
+Sodium chloride 15%
Potassium
carbonate
Sodium carbonate sat. 100
Zinc chloride
Chlorides
Aluminum chloride
Ammonium chloride 35
Potassium chloride sat.
sat.
Calcium chloride
sat.
20
Ferric chloride
20
Sodium chloride, 20%
+ Saturated Cl2
(Electrolysis solution)
Nitrates Sulfates
Seawater, Magnesium
chloride
Ammonium sulfate
Potassium sulfate
Sodium sulfate
Ammonium nitrate
Sodium nitrate
Sodium sulfite
40
60
80
100
40
60
100
24
sat. 80
60
5
Corrsion
resistance
is good for
usual salts.
Others
Hydrogen peroxide
Sodium sulfide
30
20
90
80
60
80
Potassium bichromate
Sodium sulfide
Sodium bisulfate
Wet chlorine gas
20
40
60
Sea water + Saturated Cl2
Bromine gas
Hydrogen sulfide
Sulfurous acid gas
X
X
30
6
30
5
30
5
5
30
b
20
30
10
15
43
30
10
10
30
57
25
6
5
15
5
10
28
10
25
20
20
30
100
30
b
b
65
b
b
30
b
5
35
25
20
25
25
25
sat.
25
b
b
b
b
25
25
b
60
b
30
b
X
90
X
X
X
X
X
b
30
60
90
20
20
40
5
20
40
X
40
10
30
100
5
10
20
60
X
57
30
90
30
85
20
90
90
15
30
10
10
10
10
20
90
20
90
90
5
35
25
b
b
90
25
20
25
90
90
90
20
60
5
5
5
5
10
20
10
25
20
20
30
70
30
b
b
65
b
b
30
b
5
5
5
5
10
20
10
25
20
20
30
100
30
60
b
65
b
b
30
b
93
5
35
25
20
b
b
b
b
5
35
25
20
10
25
25
sat.
25
b
b
b
b
b
b
b
60
b
30
b
25 b
sat. 60
25 b
30
b
X
90
X
Strong acid
Weak acid
X
Strong alkaliX
Weak alkali
X
20
40
20
sat.
10
20
20
24
b
b
30
b
b
b
10
24
80
90
10
10
10
90
90
90
50
20
10
b
b
30
10
30
5
90
10
90
50
20
10
b
b
30
2
60-90 X
2
60-90
15
30
10
10
10
b
b
b
30
10
b
10
10
10
X
30
b
b
b
30
95
X
95
30
b
b
30
b
b
b
X
90
24
b
b
30
b
b
b
42
20
sat.
10
20
20
Remarks
Concent’n Temp Judge
5
37
37
20
42
20
sat.
10
20
20
20
80
100
X
6
Viton
95
30
20
30-90
Compatibility
Concent’n Temp Judge Concent’n Temp Judge Concent’n Temp Judge
20
20
10
Hypochlorous Acid
10
Hydroiodic acid
Titanium
Concent’n Temp Judge
Chromic acid
Nitric acid
Hastelloy C
SUS 316
Electrode
body
material
Ryton
Seal O-ring
material
Appendix1: Chemical
8.Inquiry
Sheet
Hydrochloric acid
Ultrasonic transducer material
Electrode solution ground tip
Holder material
Polypropylene
Gases
8.
19
<8. Chemical Resistance Table >
X
Polypropylene may
sometimes be eroded
by ammonium sulfate
crystals
20
X
30
X
80
T0801.EPS
Note: "b" refers to the boiling point.
Table of Contents
TI 12B07A03-01E
1st Edition : Nov. 18,2009-00
TI12B00A20-01E First Edition March 2014
61
w w w.yo ko g a wa.c o m
20
<8. Chemical Resistance Table >
Ultrasonic transducer material
Electrode solution ground tip
Holder material
Polypropylene
Acetaldehyde
Acetone
Aniline
Appendix Chemical
Compatibility
Ether
Ethylene glycol
Ethyl alcohol
Methyl chloride
Glacial acetic acid
Glycerin
Chlorophenol
Xylene
Chlorobezene
Organic substances
Chloroform
Dioxane
Dichloroethare
Ethyl nitrate
Carbon tetrachloride
Trichloroethylene
Toluene
Benzophenone
Benzaldehyde
SUS 316
Concent'n Temp Judge
Concent'n Temp Judge
20
100 20
100 30
50 25
100 110
100
100
100
100
100
100
96
20
70
100
20
70
100
70
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
20
70
100
70
100
20
70
100
20
20
100
20
20
70
100
70
20
X
20
20
20
X
X
X
100 20
100 70
100 100
Hastelloy C
Seal O-ring
material
Titanium
Viton
Concent'n Temp Judge Concent'n Temp Judge Concent'n Temp Judge
Electrode
body
material
Ryton
100
20
100
b
100 25
100
90
100 25
100 25
100
20
100
100
90
100
20
100
20
100
20
100
90
100
90
100
90
100
100
90
90
100
100
20
90
100
90
100
90
100
100
25
90
100
90
100
20
100
100
20
20
100 25
X
b
Remarks
Concent'n Temp Judge
100 25
100 24
X
100 25
X
X
X
X
X
100
b
100
b
100
b
X
X
100
105
90
100
b
b
100
b
90
100
b
b
145
100
24
X
Benzyl alcohol
benzene
100 20
100
30
Fomaldehyde
10
10
37
b
70
100
37
b
100
30
37
b
100
25
X
X
X
Methylnaphthelen
Methyl ethyl ketone
Methyl alcohol
Nitrobenzene
Acetic acid
Phenol
Benzonic acid
Motor oil
Petroleum ether
Kerosene
Tartaric acid
Oil and fats
Carbon sulfide
100 20
70
100 20
100 20
100 70
100 100
100 20
100 70
100 100
100 20
100 100
100
100 20
100 70
100 100
100 20
100 20
100 70
10
40
10
60
10
80
100 70
100 20
100 25
X
10
b
95
30
10
b
95
30
X
95
30
101
X
50
100
100 25
X
50
100
100 180
50
100
100 180
100 25
Note: "b" refers to the boiling point.
100 25
T0802.EPS
9.2 Reference in website
For detailed information, refer to the following site;
http://www.coleparmer.com/Chemical-Resistance
X
Table of Contents
62 TI12B00A20-01E First Edition March 2014
Back to the pHuture
Notes
Appendix1: Chemical
Compatibility
Table of Contents
TI12B00A20-01E First Edition March 2014
63
w w w.yo ko g a wa.c o m
Appendix 2:
Definitions
Appendix 2: Definitions
Activity
The activity is a measure of the amount of ions
chemically active in a concentration of the ions
in a solution. This means that the participation
of ions in a chemical reaction is not only
determined by the concentration but also by
the presence of other ions in the solution. In
concentrated solutions the activity of the ions is
considerably less than the total concentration
Buffer solution
A buffer solution is a solution with a
constant, accurate fixed pH value. The use
of this solution is necessary to determine
any deviations in the measuring system and
to allow accurate calibration. The most
important feature of a buffer solution is that
the pH value remains constant when it is
diluted with water, acid or base.
Activity coefficient
The activity coefficient (f) shows the ratio
between the active concentration and the
total concentration.
f = active concentration
total concentration
Calomel
Calomel is an insolule salt (H92CI2) used as
part of a reference system in the reference
electrode.
An ions
An ions are negatively charged ions, e.g.
CI–, N03 -, OH–.
Asymmetry potential (Easy)
This potential can exist where the reference
system is nonsymmetrical, where differences
exist in the glass matrix at inner and outer
faces, and unbalanced diffusion potential
occurs. The asymmetry potential can be
compensated for during calibration.
Automatic temperature
compensation
Automatic control of the sensivity of the
analyzer to compensate for influence of
temperature on the electrode system.
Note: This temperature compensation does not
compensate for the influence of temperature
on the process itself.
Buffer capacity
The buffer capacity is expressed by quantity
of millimol equivalent of normal hydrochloric
acid required at 25°C to reduce 1 litre
of buffer by 1 pH unit, or alternatively, by
quantity of millimol of normal caustic soda
that is required at 25°C to increase 1 litre of
buffer by 1 pH unit.
Buffer powder
Buffer powder is an accurately weighted
quantity of salts which when immersed in a fixed
quantity of distilled or demineralised water
produces a buffer solution of constant value.
Concentration
The concentration of a solution is determined
by the quantity of matter dissolved per
volume - or per weight of the solution.
Concentration units
g per litre (or kg/m3) The weight in g per
litre of solution.
Grammes (g) per kilogramme (kg)
The weight in g of matter per kg of
solution.
mg per kg
The weight in mg of matter per kg of
solution.
Note: This is commonly expressed as p.p.m.
(parts per million).
µg per kg
The weight in µg of matter per kg of solution.
Note: This is commonly expressed as p.p.m.
(parts per million).
Mol per litre or molar.
The weight in grammes corresponding with
the molecular weight per litre of solution.
Mol per kg or molal
The weight in g corresponding with the
molecular weight added to 1 kg solvent.
Weight percents
The weight of a dissolved matter per 100
grammes of solution.
Diaphragm (Junction)
The diaphragm is that part of the reference
electrode which joins the reference system in
the reference electrode with process liquid.
The most common types of diaphragm in use
are: •
Porous ceramic
• Porous teflon
• Sleeve
Dilution value
The dilution value of a buffer solution is
defined as the variation of the pH value that
occurs when the solution is diluted with an
equal volume of water.
Divalent
A divalent ion is an ion with a diva-lent of
positive or negative charges.
Examples are:
AB ↔ A+B
K= [A]•[B]
[AB]
Dissociation
Dissociation is separating into positive or
negative ions.
Dissociation constant (K)
The number gives the ratio between the concentration of the separated ions in a matter and
the concentration of the unseparated matter.
Electrolyte
An electrolyte is a matter that separates
ions in an aqueous solution itself. Weak
electrolytes partly dissociate1). Strong
electrolytes dissociate almost completely.
In such a solution electrical current is
conducted in equal proportions by positive
and negative ions. This is sometimes used to
obtain low diffusion potentials.
Filling liquid
The name filling liquid is often used to
describe electrolyte.
Glass electrode
A Glass electrode is a ion-specific electrode
constructed from special types of the ionsensitive glass. Glass electrodes are available
for sodium, potassium, and hydrogen ions
measurement.
Hydration
Hydration is the uniting of ions with
molecules of water.
lonisation constant
See dissociation constant.
Ion strength
The strenght of ions in a solution is
determined by both the concentration of the
ions in the solution and the nature of these
Table of Contents
64 TI12B00A20-01E First Edition March 2014
Back to the pHuture
ions. The strength of ions determines the
activity of each ion in the solution.
In a equation, the strength of ions is:
C = concentration
Z = the square of the charge of the ion.
pH analyzer or pH meter
Standard solution
An electronic unit which converts the potential
difference between the pH electrode and the
reference electrode into a standard output signal.
See buffer solution.
Polarisation
lon-specific electrode
An ion-specific electrode is an electrode
which developes an electrical potential
proportional to the logarithm of the activity
of that ion.
Polarisation of a glass electrode takes place
when it is used with a measuring instrument
whose input circuit has too low and
impedance and draws an excessive current
through the pH sensitive membrane.
Isothermal point of intersection
The isothermal point of intersection of a glass
electrode is the theoretical intersection point
of the mY/pH graphs at different temperatures.
The selection of the electrolyte in the glass
electrode determines the position of this point.
Usually this point is. at pH 7 and 0 mV.
Polyvalent
Solubility product
The solubility product of an electrolyte is the
product of the concentration of the ions in a
saturated solution.
Example:
The solubility product of
AgCI is 1,1 x 10–10.
Strong base
Monovalent
A monovalent ion contains a single positive
or negative charge (examples: H+, Na+,
Cl– etc.).
Strong acid
Reference electrode
A reference electrode is used in combination
with an ion-specific electrode. Its function is
to generate stable potentials independent of
the composition of the liquid to be measured.
Reference system
A reference system is a combination of a
metal, an insoluble salt of this metal and
electrolyte. A reference system is used in
both the glass and reference electrode.
Nernst equation
This equation is used to determine the
potential of an ion-specific electrode.
RT
• In A
E=E0 +
nF
E = measured potential
R = gas constant (8,314 J/mol. K) T =
absolute temperature in oK
n = valency of the ion
F = Faraday number (96493 Coulomb) In =
2,303 log (logarithm)
A = ion activity
E0 = electro-chemical normal potential
Response time
Nernst factor
The selectivity constant indicates increase
in output of the ion specific electrode.
Concentration of strange ions required to
produce the same output as those to which
the electrode is specific.
The description “Nernst factor” is used to
describe the combination of constants in the
Nernst quation.
RT
8,314 x 278
x 2,303=
x 2,303 = 0,0591
nF
1 x 96493
The response time of a glass electrode indicates the time it takes to follow a step change
in pH. Usually the reponse time is defined as
the time taken to reach 63% of the final value
of a step change.
Under process conditions the response time of
the measuring system depends on a number
of factors vis. the positioning of the electrode
in the process stream, the reference electrode
used etc. and consequently, the times stated
may not always be achieved in pratice.
Selectivity constant
Sensitivity
pH electrode
An ion-specific electrode for the
measurement of the hydrogen activity in a
solution.
The sensitivity of a pH glass electrode is
the mV output voltage per pH change.
Theoretically, the sensitivity of a pH electrode
is 59,16 millivolts at 25°C.
A strong acid is one which completely
dissociates:
HCI H+ + CI–
Titration curve
The titration curve indicates the pH curve
obtained when a quantity of reagent is added
to a predetermined quantity of a reagent
necessary to neutralisation a process liquid.
Appendix 2:
Definitions
Note: T he position of the isothermal point of
intersection is important for the accuracy of
the temperature correction of the electrode
system.
A polyvalent ion is an ion with more than
two positive or negative changes, e.g. P043–,
Cr6+, Fe3+.
A strong base is one which completely
dissociates:
NaOH ↔ Na+ + OH–
Silver chloride
Silver chloride (AgCI) is an insoluble salt
used in a reference system for both glass
and reference electrodes.
Salt bridge
A salt bridge is used in applications where
the electrolyte of a reference electrode
interacts with the process. A salt bridge is a
barrier solution which is compatible with the
composition of the process liquid.
Weak base
A weak base only partly dissociates into
ions.
Example:
Cu2+ + 20H–.
Cu (OH)2
Weak acid
A weak acid only partly dissociates into
ions.
Example:
2H+ + CO32–.
H2CO3
Table of Contents
TI12B00A20-01E First Edition March 2014
65
w w w.yo ko g a wa.c o m
aPPeNDIx 3: LIQUID aPPLICaTION DaTa sHeeT
Liquid Application Data Sheet
Please place checkmarks in the appropriate boxes and fill in the necessary information in the blanks.
[ QMQI CY C
Customer Data / General
Project/Reference
Inquiry No. :
Customer
TAG :
Contact Person
Department:
City
Street:
Telephone
FAX :
e-Mail
Final Destination
Expected
Delivery
needed Documention
(Certification ...)
Quoted
Services
Month :_________ Year : __________
Commissioning:
Yes
No
Training:
Yes
No
PROCESS CONDITIONS
APPenDIx 3: LIQUID
APPLICATIOn DATA sHeeT
Industry Segment
Measuring parameter
Power
Chemical
Petrochemical
Water /Waste Water
Oil/Gas
Pulp & Paper
Pharma
Food
__________
pH
ORP
contacting SC
Ind. SC
Dissolved O2
Process Sample Name
Plant Name
Measuring Range
Application
Operation
Batch
Continuous
Monitor
Control
__________
Cleaning
Chemical
Steam
Water
Hotwater
__________
Anticipated Issues
Fouling
Poisoning
Coating
Abrasion
__________
Operating Range
Min:
Norm:
max:
Temperature
Min:
Norm:
max:
Pressure
Min:
Norm:
max:
Flowrate
Min:
Norm:
max:
Min:
Norm:
max:
Conductivity
Composition (in %):
Concentration
Solids (Types):
Organics (Types):
Inorganics (Types):
Installation Data
Installation
Indoor
Mounting type
Inline
Sensor fitting type
Insertion
Process side details
Outdoor Amb. Temp.: _________ Haz. Area Class.: __________ IP: ________
Bypass
Flow Through
T-piece
chamber
open tank
Immersion
with floating ball
w/o floating ball
Process connection ( thread/flange): _________________
Immersion/insertion length:
Atline
Retractable
manual
automatic
Mounting :
_________________
__________
Top
Side
Type cleaning system
Jet Cleaning
Brush
Chemical /Spray
Ultrasonic
__________
Wetted Material
(Fitting/O-Ring)
316SS
Titanium
PVDF/PTFE
PVC
PP
Viton
EPDM
Kalrez
Analyzer
Non Ex
Intrinsic
2 wire |
4 wire (24 - 30 VDC)
Silicon
Non Incendive | Certification: _________
4 wire (80 - 240 VAC) |
Remarks:
Table of Contents
66
closed tank
TI12B00A20-01E
First Edition
March 2014
__________
Distance to Sensor: _____
4-20 mA/Hart
PB
FF
Back to the pHuture
Notes
Appendix 3: Liquid
Application Data sheet
Table of Contents
TI12B00A20-01E First Edition March 2014
67
w w w.yo ko g a wa.c o m
Notes
Table of Contents
68 TI12B00A20-01E First Edition March 2014
Back to the pHuture
Notes
Table of Contents
TI12B00A20-01E First Edition March 2014
69
w w w.yo ko g a wa.c o m
Notes
Table of Contents
70 TI12B00A20-01E First Edition March 2014
Back to the pHuture
Notes
Table of Contents
TI12B00A20-01E First Edition March 2014
71
YOKOGAWA ELECTRIC CORPORATION
World Headquarters
9-32 Nakacho 2-chome, Musashino-shi, Tokyo 180-8750, Japan
http://www.yokogawa.com/
Represented by:
YOKOGAWA CORPORATION OF AMERICA
12530 West Airport Blvd, Sugar Land, Texas 77478, USA
http://www.yokogawa.com/us/
YOKOGAWA EUROPE B.V.
Euroweg 2, 3825 HD Amersfoort, The Netherlands
http://www.yokogawa.com/eu/
YOKOGAWA ENGINEERING ASIA PTE. LTD.
5 Bedok South Road, Singapore 469270, Singapore
http://www.yokogawa.com/sq/
YOKOGAWA CHINA CO., LTD.
3F TowerD Cartelo Crocodile Building
No. 568 West Tianshan Road, Shanghai 200335, China
http://www.yokogawa.com/cn/
YOKOGAWA MIDDLE EAST B.S.C.(C)
P.O. Box 10070, Manama
Building 577, Road 2516, Busaiteen 225, Muharraq, Bahrain
http://www.yokogawa.com/bh/
YOKOGAWA INDIA LTD.
Plot No. 96, Electronic City Complex, Hosur Road, Bangalore - 560 100, India
http://www.yokogawa.com/in/
The contents of this document are subject to change without prior notice.
All rights reserved. Copyright © 2014 by Yokogawa Electric Coporation
TI12B00A20-01E
Printed in India[Ed: 1/a]