Mettler Toledo pH Theory and Guide
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
pH Theory Guide
Mettler-Toledo AG
Process Analytics
Industry
Environment
A Guide to pH Measurement –
the Theory and Practice of pH Applications
A Guide to pH Measurement –
the Theory and Practice of pH Applications
Copyright © 2013 by Mettler-Toledo AG
CH-8902 Urdorf/Switzerland
Content
Preface
10
1 Introduction to pH
1.1 Acidic or alkaline?
1.2 Why are pH values measured?
1.3 The tools for pH measurements
1.3.1 The pH electrode
1.3.2 Reference electrodes
1.3.3 Combination electrodes
1.4 What is a pH measuring system?
11
11
13
14
15
17
18
19
2 Practical considerations
2.1 The pH measuring system
2.2 Obtaining an accurate pH measurement
2.2.1 General principles of pH measurement
2.2.2 Industrial pH measurement
2.2.3 Signal processing and environmental influences
2.2.4 Calibration
2.2.5 Buffer solutions
2.3 How to maintain a reliable signal
2.3.1 Maintenance of the electrode function
2.3.2 Storage
2.3.3 Temperature compensation
2.4 Troubleshooting
2.4.1 Instructions and comments for the trouble
shooting diagram
20
20
21
21
23
26
30
31
32
32
35
35
39
3 Intelligent Sensor Management
3.1 Signal integrity
3.2 Pre-calibration
3.3 Predictive diagnostics
3.4 Asset management software
3.4.1 Electronic documentation
3.4.2 Sensor management
44
44
45
45
47
47
48
39
pH Theory Guide
METTLER TOLEDO
5
4 Electrode selection and handling
4.1 Different kinds of junction
4.1.1 Ceramic junctions
4.1.2 PTFE annular diaphragm
4.1.3 Open junctions
4.1.4 Dual-membrane without junction
4.2 Reference systems and e lectrolytes
4.3 Types of membrane glass and membrane shape
4.4 pH e lectrodes for specific applications
4.4.1 Highly accurate problem solver
4.4.2 Complex samples or such of unknown
composition
4.4.3 Semi-solid or solid samples
4.4.4 At the toughest applications in chemical
process industries
4.4.5 Prepressurized electrolyte pH electrodes
4.4.6 Dual-membrane pH electrodes
4.4.7 pH measurements in high purity water samples
4.4.8 Installation in an upside-down position
4.4.9 Non-Glass (ISFET) pH electrodes
4.4.10 For low maintenance and simple installation
4.5 Electrode maintenance
4.6 Electrode storage
4.6.1 Short term storage
4.6.2 Long term storage
4.7 Electrode cleaning
4.7.1 Blockage with silver sulfide (Ag2S)
4.7.2 Blockage with silver chloride (AgCl)
4.7.3 Blockage with proteins
4.7.4 Other junction blockages
4.8 Electrode regeneration and lifetime
6
pH Theory Guide
METTLER TOLEDO
50
50
50
52
53
53
54
58
59
59
60
61
61
62
63
64
65
65
66
67
67
68
68
68
69
69
69
69
69
5 Comprehensive pH theory
5.1 Definition of the pH value
5.2 Correlation of concentration and activity
5.3 Buffer solutions
5.3.1 Buffer capacity (ß)
5.3.2 Dilution value (∆pH)
5.3.3 Temperature effect (∆pH / ∆T)
5.4 The measurement chain in the pH measurement setup
5.4.1 pH electrode
5.4.2 Reference electrode
5.5 Calibration / adjustment of the pH measurement setup
5.6 The influence of temperature on pH measurements
5.6.1 Temperature dependence of the electrode
5.6.2 Isothermal intersection
5.6.3 Further temperature phenomena
5.6.4 Temperature dependence of the measured
sample
5.7 Phenomena in the case of special measuring solutions
5.7.1 Alkaline error
5.7.2 Acid error
5.7.3 Reactions with the reference electrolyte
5.7.4 Organic media
5.8 Signal processing
71
71
72
75
77
78
78
78
80
81
84
85
85
86
87
6 Mathematical parameters
99
88
89
89
90
90
91
93
pH Theory Guide
METTLER TOLEDO
7
Figures
Figure 1
Figure 2
Figure 3
Figure 4
Figure 5
Figure 6
Figure 7
Figure 8
Figure 9
Figure 10
Figure 11
Figure 12
Figure 13
Figure 14
Figure 15
Figure 16
Figure 17
Figure 18
Figure 19
Figure 20
Figure 21
Figure 22
Figure 23
Figure 24
Figure 25
Figure 26
Figure 27
Figure 28
8
pH Theory Guide
METTLER TOLEDO
The reaction of an acid and a base forms water
Dissociation of acetic acid
The formula for calculating the pH value from
the concentration of hydronium ions
pH values for some chemicals and everyday products
The reaction of ammonia with water
The relationship between the amount of acid in solution
and the output potential of a pH electrode
The measurement assembly of pH and reference
electrode
Cross sections through the glass membrane
pH electrode with pH-sensitive membrane
Reference electrode with reference electrolyte,
reference element and junction
Typical combination pH electrode with inner pH sensor
and outer reference element
pH measurement system
InTrac 776 e
Industrial measuring sites
Signal transformation
Complete measurement system
Electrode with built-in electrolyte bridge
Calibration line and isothermal intersection points
Symmetrical structure of an Equithal®-system in
comparison with a conventional electrode
Troubleshooting diagram
Electrode with ceramic junction
Example of electrode with PTFE diaphragm
Example of electrode with open junction
Dual-membrane pH electrode
Schematic drawing of the ARGENTHAL™ reference
system
Differently shaped pH membranes
InPro 200x (i)
InPro 426x (i)
11
11
11
12
12
14
15
16
16
17
18
19
24
25
26
29
34
37
38
40
51
52
53
54
55
58
60
60
Figure 29
Figure 30
Figure 31
Figure 32
Figure 33
Figure 34
Figure 35
Figure 36
Figure 37
Figure 38
Figure 39
Figure 40
Figure 41
Figure 42
Figure 43
Figure 44
Figure 45
Figure 46
Figure 47
Puncture pH electrode
61
InPro 480x (i)
62
InPro 325x (i)
63
InPro 4850 i
64
pHure Sensor™
64
InPro 3100 (i)
65
InPro 3300 (ISFET pH sensor)
66
InPro 4501
66
InPro 4550
67
Buffering capacity of acetic acid
77
Temperature dependence for the pH electrode slope
factor
79
Different sources of potential in a combination electrode 79
Ion mobility and diffusion of ions through a junction
82
Left: offset adjustment of a pH electrode in the pH
meter, right: slope adjustment of a pH electrode.
Solid lines show ideal behavior, dashed lines show
real behavior
85
Isothermal intersection, theory and practice
87
Illustration of alkaline and acid error electrode behavior 90
pH scale for different solvents
92
Typical process control loop
93
Intersection of the process control system and
sensor / activator s ystem
94
pH Theory Guide
METTLER TOLEDO
9
pH Theory Guide
Preface
The aim of this book is to give a representative description of pH meas
urement in the process industries. The actual sensor, the pH electrode,
is therefore the main focus of the text. Correct sensor use is fundamental for a meaningful pH measurement. Accordingly, both practical and
theoretical requirements are discussed in depth so that the measuring
principle is understood and an accurate measurement made possible.
The first section (practical considerations) of the book describes the
sensor, and the other elements that constitute a pH measurement system. Together with a troubleshooting diagram, this section gives the information needed in order to ensure the correct working of the pH electrodes for long periods of time. The second, application orientated
section gives solutions to different measuring tasks, giving examples
from the lab and from industry. The last, theoretical part explains the
basis of the pH measurement and completes, by further explanation,
the information given in the first section.
In addition, this book is outlined to be a useful tool in solving different
measuring tasks. Thereby it can be read either in its totality or in parts.
Urdorf, Switzerland, January 2013
10
pH Theory Guide
METTLER TOLEDO
1
Introduction to pH
1.1
Acidic or alkaline?
Why do we classify an everyday liquid like vinegar as being acidic?
The reason is that vinegar contains an excess of hydronium ions
(H3O+) and this excess of hydronium ions in a solution makes it acidic.
An excess of hydroxyl ions (OH–) on the other hand makes something
basic or alkaline. In pure water the hydroniumn ions are neutralized by
hydroxyl ions, therefore this solution has a neutral pH value.
H3O+ + OH– ↔ 2 H2O
Figure 1
The reaction of an acid and a base forms water.
If the molecules of a substance release hydrogen ions or protons
through dissociation we call this substance an acid and the solution
becomes acidic. Some of the most well-known acids are hydrochloric
acid, sulfuric acid and acetic acid or vinegar. The dissociation of
acetic acid is shown below:
CH3COOH + H2O ↔ CH3COO– + H3O+
Figure 2
Dissociation of acetic acid.
Not every acid is equally strong. Exactly how acidic something is, is
determined by the total number of hydrogen ions in the solution. The
pH value is then defined as the negative logarithm of the hydrogen ion
concentration. (To be precise, it is determined by the activity of the hydrogen ions. See “5.2 Correlation of concentration and activity“ on
page 72 for more information on the activity of hydrogen ions).
pH = – log [aH+]
Figure 3
The formula for calculating the pH value from the concentration of
hydronium ions.
pH Theory Guide
METTLER TOLEDO
11
pH Theory Guide
The quantitative difference between acidic and alkaline substances
can be determined by performing pH value measurements. A few
examples of pH values of everyday substances and chemicals are
given in Figure 4 below.
Food & Beverages / Household products
Orange juice
Cheese
Coca Cola
Beer
Lemon juice
0
1
2
Egg white
3
4
Water
Antacid Mg(CH)2
Milk
5
6
Borax
7
8
Acetic acid
Sulfuric
0.6 % (0.1M)
acid 4.9 %
(1M)
Hydrochloric acid Hydrocyanic acid
0.37 % (0.1M)
0.27 % (0.1M)
9
10
11
12
13 14
Caustic
soda 4 %
Calcium
carbonate (sat)
Ammonia sol. 1.7 % (1M)
Ammonia sol. 0.017 % (0.01M)
Potassium acetate 0.98 % (0.1M)
Sodium hydrogen carbonate 0.84 % (0.1M)
Chemicals
Figure 4
pH values for some chemicals and everyday products.
The alkaline end of the scale is between pH 7 and 14. At this end of the
scale the hydroxyl or OH– ions are present in excess. Solutions with
these pH values are created by dissolving a base in an aqueous solution. The base dissociates to release hydroxyl ions and these make the
solution alkaline. Among the most common bases are sodium hydroxide, ammonia, and carbonate.
NH3 + H2O ↔ NH4+ + OH–
Figure 5
The reaction of ammonia with water.
The whole scale of pH values in aqueous solutions includes both the
acidic and alkaline ranges. The values can vary from 0 to 14, where
pH values from 0 to 7 are called acidic and pH values from 7 to 14 are
termed alkaline. The pH value of 7 is neutral.
12
pH Theory Guide
METTLER TOLEDO
1.2
Why are pH values measured?
We measure pH for a lot of different reasons, such as:
• to produce products with defined properties – During production it
is important to control the pH to ensure that the end product conforms with the desired specifications. The pH can dramatically alter
the properties of an end product such as appearance or taste.
• to lower production costs – This is related to the above mentioned
reason. If the yield of a certain production process is higher at a
given pH, it follows that the costs of production are lower at this pH.
• to avoid doing harm to people, materials and the environment –
Some products can be harmful at a specific pH. We have to be careful not to release these products into the environment where they can
be a danger to people or damage equipment. To be able to determine
whether such a substance is dangerous we first have to measure its
pH value.
• to fulfill regulatory requirements – As seen above, some products
can be harmful. Governments therefore put regulatory requirements
in place to protect the population from any damage caused by dangerous materials.
• to protect equipment – Production equipment that comes into contact with reactants during the production process can be corroded by
the reactants if the pH value is not within certain limits. Corrosion
shortens the lifetime of the production line, therefore monitoring pH
values is important to protect the production line from unnecessary
damage.
• for research and development – The pH value is also an important
parameter for research purposes such as the study of biochemical
processes.
These examples describe the importance of pH in a wide range of
applications demonstrating why it is so often determined.
pH Theory Guide
METTLER TOLEDO
13
pH Theory Guide
1.3
The tools for pH measurements
To be able to measure pH you need to have a measurement tool which
is sensitive to the hydrogen ions that define the pH value. The principle
of the measurement is that you take a sensor with a glass membrane
which is sensitive to hydrogen ions and observe the reaction between it
and a sample solution. However, the observed potential of the pH-sensitive electrode alone does not provide enough information and so we
need a second sensor. This is the sensor that supplies the reference
signal or potential for the pH sensor. It is necessary to use the difference between both these electrodes in order to determine the pH value
of the measured solution.
The response of the pH-sensitive electrode is dependent on the H+ ion
concentration and therefore gives a signal that is determined by how
acidic / alkaline the solution is.
The reference electrode on the other hand is not responsive to the H+
ion concentration in the sample solution and will therefore always produce the same, constant potential against which the pH sensor potential is measured.
The potential between the two electrodes is therefore a measure of
the number of hydrogen ions in the solution, which by definition gives
the pH value of the solution. This potential is a linear function of the
hydrogen concentration in the solution, which allows quantitative
measurements to be made. The formula for this function is given in
Figure 6 below:
RT
log [aH+]
E = E0 + 2.3
nF
Figure 6
The relationship between the amount of acid in solution and the output
potential of a pH electrode.
E = measured potential
R = gas constant
n = ionic charge
14
pH Theory Guide
METTLER TOLEDO
E0 = constant
T = temperature in degrees Kelvin
F = Faraday constant
Glass
electrode
Figure 7
High
impedance
pH meter
Reference
electrode
The measurement assembly of pH and reference electrode.
In Figure 7 a pH measurement setup with two separate electrodes, a
pH electrode and a reference electrode, is shown. Nowadays, a merger
of the two separate electrodes into one sensor is very common and
this combination of reference and pH electrodes is called the combination pH electrode. Each of these three electrodes is different and has its
own important features and properties.
1.3.1
The pH electrode
The pH electrode is the part that actually senses the pH in the solution.
It consists of a glass shaft with a thin glass membrane at the end,
sensitive to H+ ions. The outside of this membrane glass forms a gel
layer when the membrane comes into contact with an aqueous solution. A similar gel layer is also formed on the inside of the membrane
glass, since the electrode is filled with an inner aqueous electrolyte
solution. An example of this gel layer is shown in Figure 8 below:
pH Theory Guide
METTLER TOLEDO
15
Glass
membrane
▲
Internal
buffer
SiO3
SiO3
Li+
+
internal buffer
Li+ SiO3+
SiO3 Li
Li
SiO3 SiO3
Li+ Li+
SiO3 SiO
Li+
SiO3
3
SiO3 + SiO Li+
Li
3
▼
▼
▼
Positive
charge
▼
▲
▼
H+
Outer
gel layer
Negative
charge
▲
Measured
solution
H+
▲
pH Theory Guide
Inner
Inner buffer
H+= constant gel layer
H+ Acidic solution
Alkaline solution
H+
Glass membrane (0.2–0.5 mm)
Gel layer ca. 1000 A (10-4 mm)
Figure 8
Cross sections through the glass membrane.
The H+ ions in and around the gel layer can either diffuse into or out
of this layer, depending on the pH value and therefore the H+ ion concentration of the measured solution. If the solution is alkaline the H+
ions diffuse out of the layer and a negative charge is established on
the outer side of the membrane. If the solution is acidic the reverse
happens, H+ ions diffuse into the layer and a positive charge builds-up
on the outer side of the membrane. Since the glass electrode has an
internal buffer with a constant pH value, the potential on the inner surface of the membrane remains constant during the measurement. The
pH electrode potential is therefore the difference between the inner and
outer charge of the membrane. A drawing of a standard pH electrode is
shown in Figure 9 below.
Inner
buffer
Membrane
Figure 9
16
pH Theory Guide
METTLER TOLEDO
Lead-off
element
Shield
pH electrode with pH-sensitive membrane.
Socket
1.3.2
Reference electrodes
The purpose of the reference electrode is to provide a defined stable
reference potential for the pH sensor potential to be measured against.
To be able to do this the reference electrode needs to be made of a
glass which is not sensitive to the H+ ions in the solution. It must also
be open to the sample environment into which it is dipped. To achieve
this, an opening or junction is made in the shaft of the reference electrode through which the inner solution or reference electrolyte is in contact with the sample. The reference electrode and pH half-cell have to
be in the same solution for correct measurements. A picture of a typical
reference electrode is shown in Figure 10 below:
Junction
Reference
Element
Electrolyte
Refill
opening
(option)
Figure 10 Reference electrode with reference electrolyte, reference element and
junction.
The construction of the electrode is such that the internal reference
element is immersed in a defined reference buffer and is indirectly in
contact with the sample solution via the junction. This contact chain
ensures a stable potential.
There are several reference systems available, but the one used almost
exclusively today is the silver / silver chloride system. The potential
of this reference system is defined by the reference electrolyte and the
silver / silver chloride reference element. It is important that the reference
electrolyte has a high ion concentration which results in a low electrical resistance (see “5.4 The measurement chain in the pH measurement setup“ on page 78 for more details).
pH Theory Guide
METTLER TOLEDO
17
pH Theory Guide
Since the reference electrolyte flows into the sample solution during
measurement, you should be aware of any possible reactions between
the reference electrolyte and the sample solution as this can affect the
electrode and measurement.
1.3.3
Combination electrodes
Combination electrodes (see Figure 11 below) are much easier to handle than two separate electrodes and are very commonly used today.
In the combination electrode the pH-sensitive glass electrode is concentrically surrounded by the reference electrode filled with reference
electrolyte.
The separate pH and reference parts of the combination electrode have
the same properties as the separate electrodes; the only difference is
that they are combined into one electrode for ease of use. Only when
the two components of the combination electrode are expected to have
very different life expectancies is the use of individual pH and reference
electrodes recommended rather than a single combined electrode.
To further simplify pH measurements, it is possible to house a tem
perature sensor in the same body as the pH and reference elements.
This allows temperature compensated measurements to be made.
Such electrodes are also called 3-in-1 electrodes.
Membrane
Inner
buffer
Junction
Reference
electrolyte
Lead-off
element
Socket
Reference
element
Figure 11 Typical combination pH electrode with inner pH sensor and outer reference
element.
18
pH Theory Guide
METTLER TOLEDO
1.4
What is a pH measuring system?
An electrode housing is necessary in order to protect and securely hold
the pH electrode in a continuous industrial process.
The function of a pH transmitter is to present the signals of the electrode in a suitable way; for instance with the help of a pH display or an
output for a recording device. The different components of the pH measuring system can be summarized as follows:
pH transmitter
Cable
Measuring system
Combination pH electrode
pH electrode
assembly
Electrode housing
Figure 12 pH measurement system.
pH Theory Guide
METTLER TOLEDO
19
pH Theory Guide
2
Practical considerations
2.1
The pH m
easuring system
Correct pH measurement can only be achieved through the proper
design of the measuring system, including the measuring site.
What makes a pH measuring system?
In the lab:
Suitable electrode
Holder
Positions and protects the pH electrode against
mechanical damages
Thermometer
Temperature control
Laboratory pH meter
Calibration of electrode, pH value display
Buffer solution
Calibration
Electrolyte solution
Storage and refill of electrodes
Dist. water
Cleaning of the electrode
Stirrer
Generates a homogenous measuring solution
(medium)
Beaker
Measuring, calibrating and cleaning
In industry:
Suitable electrode
20
pH Theory Guide
METTLER TOLEDO
Electrode housing
Protection. It should be of such a design that measurement and maintenance of the electrode can be
optimally done
Temp. sensor
Temperature compensation
pH transmitter
Calibration, monitoring and control of the process
Buffer solution
Calibration
Electrolyte solution
Storage and refill of electrolyte
Cleaning system
Cleaning and calibrating of the electrode
2.2
Obtaining an accurate pH measurement
2.2.1
General principles of pH measurement
For an optimal pH measurement to be possible, the correct electrode
must first be selected.
This is done with respect to the following criteria:
• chemical composition
• temperature
• pH range
• pressure
• vessel size
Special solutions demand special electrolytes
Under certain circumstances, in special applications, it is necessary
not only to chose the right electrode, but also to use a special electrolyte. This is, for instance, the case with very protein-rich solutions, or
with non-aqueous or partly aqueous solutions.
Modern pH transmitters allow both manual and automatic temperature
compensation (see “5.6 The influence of temperature on pH measurements“ on page 85). If the pH measurement is, however, always taken
at the same temperature, an automatic temperature compensation is
unnecessary.
To document the measured values it is advisable to use a transmitter
with a recorder output.
– Defined measuring conditions
The requirements for reproducible and accurate pH measurements are
to have defined measuring conditions. Two main points have to be
considered:
• Known temperature: If pH values of different measurements are to
be compared, it is important that they were made at the same temperature.
pH Theory Guide
METTLER TOLEDO
21
pH Theory Guide
• Homogeneity: A solution in which sediments are present has to be
stirred in order to stay physically and chemically homogeneous.
Time can also play an important role, especially with samples of a
low buffer capacity, in samples with a low (or no) salt content. CO2,
(from the atmosphere), for example, dissolves in such samples,
causing a change in pH value.
– Response time and accuracy of the electrode
In order to achieve accurate measuring results, calibration must be
done before each measurement. A new electrode in a standard buffer
(pH values 4; 7; 9.2) has a response time of less than 5 sec. to
achieve a stable reading to ± 0.01 pH units.
At extreme pH values the response time of the electrode may increase.
If a stable pH value is not reached over a longer period of time, the
cause may be one of a variety of problems (see “2.4 Troubleshooting“
on page 39).
The accuracy of the measured pH value depends on the maintenance
of the electrode, the measuring solution (extreme pH values, contamination) the temperature, the pressure, the choice of the electrode and
fresh buffer solution. A standard buffer solution should have an accuracy of ± 0.02 pH units.
– How to handle the electrode
The electrode is a sensitive device and should not be misused (for example, to stir the measured solution). The electrode should be attached
to an electrode holder.
To clean the electrode, it should only be rinsed (with distilled water
or buffer solution) and carefully dabbed dry with a clean tissue.
The electrode should not be rubbed, as this could give rise to electrostatic forces, which increase the response time of the electrode
(see “2.3 How to maintain a reliable signal“ on page 32).
22
pH Theory Guide
METTLER TOLEDO
2.2.2
Industrial pH measurement
The pH electrode assembly should be mounted in a place where it
is easily reached and where defined conditions exists. Unlike laboratory measurements, industrial measurements are mainly continuous
(on-line), see “Figure 14 Industrial measuring sites.“ on page 25. Usually
the process occurs in closed stirring vessels (chemical reactors, bioreactors). Measurements may, however, also be taken in pipe-lines,
basins, or open canals. The process conditions may be very different.
In addition to high temperatures and pressures, an electrical current
may also have to be considered. An industrial pH electrode assembly
consisting of a pH electrode and an electrode housing must be adapted
to these circumstances. For industrial processes efficiency is critical.
Hence, simple maintenance is of main interest.
The choice of a suitable pH electrode assembly is determined by the
installation possibilities. Ideally, a pH electrode assembly should be
mounted in a place that is easily reached and where representative
and defined conditions of the process exist. In a stirred vessel the electrode assembly is often installed laterally with the aid of a weld-in
socket. Flange joints or screw connections are also used. In order to
ensure that the glass membrane is always filled with the inner buffer
an installation angle of at least 15° above horizontal should be used.
It is difficult to install side mounted lateral support fittings in vessels
with an inner lining of glass, enamel or rubber. In these vessels pH
measurements must therefore be made through the cover with very
long electrode assemblies.
Especially in vessels containing solid components the stirring forces
have to be considered. A basic principle is to make sure that the pH
electrode assembly remains immersed in the solution, independent of
the level of the solution. If there are no more free ports available in the
cover, the pH value may be measured in a bypass. In this measuring
arrangement you have to take a delayed change of the pH value into
account. The installation of the bypass system must be done in such a
way that clogging does not occur.
pH Theory Guide
METTLER TOLEDO
23
pH Theory Guide
A retractable housing has great advantages when installed literally in a
pipe-line. The pH electrode may be checked, cleaned, calibrated and
replaced without interrupting the process. When installed in a pipe-line,
no bypass valves are necessary. The possibility of withdrawing the
electrode during the process helps to ensure a correct pH measurement. Retractable pH housings meet the requirements for an automatic
pH measuring system. Probes without a retractable function enable the
maintenance of an electrode only in combination with an interruption of
the process, emptying of the vessel, etc. in pressurized systems.
After having clarified the installation possibilities, the process
determines the choice of suitable electrode and housing. Both components of the pH electrode assembly should suit the process requirements.
Electrodes with a liquid electrolyte must be operated with an overpressure, preventing the measuring solution from entering into the
reference electrolyte. The overpressure also ensures a constant
cleaning of the junction. Housings for liquid electrolyte electrodes
therefore posses a pressure chamber and a pressure indicator (see
Figure 13 below). Using this type of pH electrode assembly requires
control of the overpressure and the periodic inspection of the level of
the reference electrolyte. Because of their ability to be refilled, liquid
filled electrodes have a long life expectancy if maintained correctly.
Also, in the case of difficult measurements a special electrolyte may
be used.
Figure 13 InTrac 776 e
24
pH Theory Guide
METTLER TOLEDO
Electrodes with gel or polymer electrolytes require very little maintenance. Checking the overpressure and the electrolyte level is not necessary. Polymer electrolyte electrodes are pressure resistant and do not
need a classical junction due to their special construction; therefore,
they are less sensitive to contamination. Electrodes with a gel electrolyte may be supplied with an overpressure. Since their life expectancy
is generally good, a gel or polymer electrode is often the most economical alternative due to their low maintenance costs. In addition, the
probe can be kept small and simple, since no pressure chamber is
needed.
The probe in the probe housing has to have the same corrosion resistance as the reactor. In addition to stainless steel, alloyed materials
(e.g. Hastelloy) and plastics (e.g. PVDF) are used.
1
1 Top-entry system in closed or pressurized
reactor/vessel
2 Side-entry system in reactor or vessel
3 Pipe-entry or direct-mounted system
4 Flow-through system
2 5 Immersion system in pressure-free, open
basin or vessel
6 Off-line measurement system (not illustrated)
3
4
5
Stationary or direct-mounted system
Retractable system
(electrode can be replaced during an on-going process)
Figure 14 Industrial measuring sites.
pH Theory Guide
METTLER TOLEDO
25
pH Theory Guide
2.2.3
Signal processing and environmental influences
The combination electrode provides a potential which is specific for the
pH value. This signal has a high resistance. In order to represent it as a
pH value, a special transmitter is needed (see Figure 15 below).
The potential of the pH electrode is also influenced by the temperature.
Therefore, the pH transmitter offers the possibility of manual or automatic temperature compensation (see “5.6 The influence of temperature on pH measurements“ on page 85).
Electrode
potential E
pH value
Temperature t
pH transmitter
Calibration data
Figure 15 Signal transformation.
26
pH Theory Guide
METTLER TOLEDO
Calibration and control
The electrode is characterized by its zero point and its slope. In addition to the two-point calibration (see “2.2.4 Calibration“ on page 30),
other methods exist which can monitor the performance of the electrode. For instance, a so called one-point calibration may be performed, provided that the slope remains constant.
If a process sample is taken instead of using the buffer solution, and
its pH value is measured in the lab, the electrode does not have to be
removed from the process. The pH transmitter can be set using the pH
value obtained for the sample in the laboratory. However, it has to be
taken into consideration that the pH value of the measured solution is a
function of different parameters such as temperature, dissolved C02
level, completion of reaction, etc.
A second possibility for controlling the system is to determine both the
zero point and the slope in the laboratory and subsequently enter the
values into the pH transmitter.
Temperature compensation
Automatic temperature compensation is generally used in industry,
except when both calibration and pH measurement occur at a constant
temperature (see “2.2.3 Signal processing and environmental influences“ on page 26). If the process temperature does not differ from room
temperature by more than 10 °C, temperature compensation is often
not necessary, since the measuring error is less than 0.15 pH units
(between pH 3 and 11; see “5.6 The influence of temperature on pH
measurements“ on page 85).
– Accuracy and reproducibility of measuring values
The main differences between pH measurement in the laboratory and
in industry are the frequency of calibration and the accessibility of the
pH assembly. The accuracy and reproducibility of the measured values
depend basically on the frequency of calibration and maintenance of
the electrode.
pH Theory Guide
METTLER TOLEDO
27
pH Theory Guide
– Environment
Environmental influences act mainly on the cable and on the pH transmitter. The most important factors are the surrounding temperature,
relative humidity, and electromagnetic fields. Hence, in industry protection class IP65 should be observed, as well as a robust design. In
industry different hazardous areas may exist. The pH transmitter has to
be designed according to the requirements for each safety zone.
– Cable
Since the signals are high impedance, long cables are susceptible to
disturbances. At a cable length of over 10 meters a preamplifier is desirable; at a length of over 20 meters, indispensable.
A coaxial cable is used as an electrode connection cable. In the case
of strong electromagnetic disturbances (caused, for example, by motors) it is wise to use a triaxial cable ( a coaxial cable with an additional shield), for which a correct grounding of the outer shield is
needed. The outer shield should be grounded only in one place.
Signal interpretation
For signal conditioning, transmitters are galvanically isolated between
input and output to prevent ground loops. This helps to avoid disturbances caused by potential differences between the transmitter ground
and the ground potentials in the vessel, reactor, or in the pipe-line.
These ground currents will otherwise flow through the reference electrode and appear as a signal error or even destroy the reference
system.
28
pH Theory Guide
METTLER TOLEDO
These kinds of disturbances are eliminated without exception through
galvanic isolation.
PLC
Transmitter
Electrode in housing
Measured solution
Figure 16 Complete measurement system.
Digital measurement systems
The development over the past decade of digital process analytical
measurement systems for pH and other parameters, such as METTLER
TOLEDO‘s Intelligent Sensor Management (ISM) technology, has brought
about significant improvements to sensor operations. These include
digital transmission of the measurement signal from the sensor to the
transmitter. This eliminates the influence of the aforementioned environmental factors and cable length on signal integrity (see “3 Intelligent
Sensor Management“ on page 44).
pH Theory Guide
METTLER TOLEDO
29
pH Theory Guide
2.2.4
Calibration
Both the zero point, the point where the pH electrode delivers 0 mVpotential and the slope of the calibration line show manufacturing
dependent tolerances and will change after exposure to the measuring
solutions.
Therefore the pH electrode has to be calibrated with accurately defined
buffer solutions. In order to make an exact calibration, the zero point of
the measuring chain has to be known. This is generally at pH 7 (see
“5.5 Calibration / adjustment of the pH measurement setup“ on page 84).
Two buffer solutions chosen according to the desired accuracy and
measuring range should be used. The value of the first buffer solution
should be close to zero mV (at pH 7). The second buffer solution
should have a pH value within the measuring range.
It should be noted that the two buffer solutions must have a difference
in pH values of at least two pH units. As already mentioned, the pH
value is temperature dependent. Therefore, it is important that the temperature curve of the buffer is known. Additionally, the buffer and the
electrode must be at the same temperature, or you must wait for the
temperature to reach equilibrium.
The zero point calibration (pH 7, first buffer) always has to take place
before the slope calibration (with a pH value close to the measuring
value, second buffer) except when working with a microprocessor controlled system.
If a very accurate measurement is required, it is advisable to repeat the
zero point calibration after the slope calibration.
30
pH Theory Guide
METTLER TOLEDO
The following factors directly influence the accuracy of the
calibration as well as the pH measurement:
• Buffer solutions
• Temperature measurement and temperature compensation
• Condition of the junction and the reference system (contamination,
etc.)
• Working technique
The stability of the zero point and the slope depends on the composition of the measuring solution as well as on the temperature. It makes
little sense to make general statements regarding the calibration frequency. Therefore, when working with unknown solutions it is advisable to repeat the calibration often in the beginning.
When the calibration values are stable, the calibration frequency may
be extended. In general the calibration frequency depends on the desired accuracy. To be able to calibrate during the process, an electrode
in a retractable housing is very useful.
2.2.5
Buffer solutions
For extremely accurate measurements National Institute of Standards
and Technology (NIST) buffer solutions are recommended. These buffer solutions are the basis of the practical pH scale and have been
adapted in the DIN 19266. However, they are almost exclusively used
in the laboratory. Buffers customarily used (e.g. potassium dihydrogen
phosphate and disodium hydrogen phosphate, borax or sodium carbonate) are distinguished by their high buffer capacity and long-term
stability. These buffer solutions are used in most cases and should
have an accuracy of not less than 0.02 pH units. The temperature
curve of the buffer solution must be known.
Buffer solutions last for a limited period of time. (Sealed bottles last for
about one year when stored correctly). Using good lab practice should
prevent buffer contamination. Carbon dioxide from the air may contaminate buffer solutions with a high pH value.
pH Theory Guide
METTLER TOLEDO
31
pH Theory Guide
Most modern transmitters will feature stored temperature curves for a
range of buffers.
2.3
How to maintain a reliable signal
The pH electrode is an electrochemical sensor whose efficiency depends on a reversible interaction between the sensor and the measuring solution.
The accuracy of the signal is decreased when residue on the glass
membrane or reactions of the reference system disturb this exchange.
2.3.1
Maintenance of the electrode function
An increase in response time of the electrode, a decrease of the slope,
or a zero point shift, are all phenomena due to either a reaction with
the measuring solution (contamination) or to the ageing of the
electrode.
Every electrode ages as a result of the chemistry of the glass even
when it is not used for measurements. High temperatures increase this
ageing process. Under laboratory conditions a life span of up to three
years can be expected, With continuous measurements at 80 °C, the
life span of an electrode could be significantly decreased (perhaps to a
few months).
If a reaction between the measuring solution and the electrolyte causes
disturbances, using an electrode with a silver-ion trap reference system, an e lectrolyte bridge, and / or a special electrolyte often improves
conditions (see Figure 17 on page 34).
32
pH Theory Guide
METTLER TOLEDO
How to avoid contamination:
• Periodic rinsing of the electrode with a suitable solvent.
• If the risk exists that solids may be deposited on the surface of the
membrane they can be removed by increasing the stirring or flow
velocity respectively.
The electrode should be cleaned when the junction or the glass membrane is contaminated. Depending on the type of contamination, different cleaning methods are recommended:
Measuring solutions containing
proteins: (contamination of the
junction)
The electrode is soaked in
pepsin / HCI for several hours
Measuring solutions containing
sulfides (black junction)
The junction is soaked in urea / HCI
solution until bleached
Lipid and other organic measuring
solutions
Short rinsing of the electrode with
acetone and ethanol
Acid and alkaline soluble contaminations
Einsing the electrode with
0.1 mol / L NaOH or 0.1 mol / L HCI
for a few minutes
After these treatments the electrode has to be soaked in the reference
electrolyte for 15 minutes. Also a recalibration has to be done before
any new measurements can be carried out, since the cleaning solution
diffuses into the junction during the cleaning procedure, and may
cause diffusion potentials.
The electrode should only be rinsed and never rubbed or otherwise
mechanically cleaned, since this would lead to electrostatic charges.
This could cause an increase in the response time.
pH Theory Guide
METTLER TOLEDO
33
pH Theory Guide
Reference
element
Bridge
electrolyte
Reference
electrolyte
Filling port for the
bridge electrolyte
Filling port for the
reference electrolyte
Figure 17 Electrode with built-in electrolyte bridge.
Refill of electrolyte
The reference electrolyte has to be refilled or changed when:
• the conducting element of the reference electrode is no longer completely immersed in the electrolyte (because of the diffusion of the
electrolyte through the junction).
• the reference electrolyte is contaminated (because of intrusion of the
measuring solution).
• the concentration of the reference electrolyte has increased through
water evaporation.
Please note that only the electrolyte referred to on the electrode
should be used.
Pressure compensation for liquid filled electrodes
When measuring in a vessel or in a pipeline, the reference electrolyte is
kept under a slight overpressure in order to avoid contamination by the
measured solution. This procedure is only possible (and necessary)
when using an electrode with a liquid electrolyte.
34
pH Theory Guide
METTLER TOLEDO
2.3.2
Storage
Electrodes should always be stored in the reference electrolyte. This allows immediate use of the electrode and ensures a short response
time.
When stored dry for long periods, many electrodes must be reactivated
by soaking for several hours before use in order to get optimal measuring results. If these measures are not sufficient the electrode may be
made functionable by treating it with a special reactivation solution followed by subsequent conditioning in the reference electrolyte (e.g. over
night). If stored in distilled water the electrode will have a longer response time.
2.3.3
Temperature compensation
The pH range (0 -14) is determined through the ion product of the
water, which to a very small extent dissociates into H+ and OH– ions.
[H+] [OH–] = 10 –14 = I (25 °C)
The ion product I is strongly dependent on temperature.
The temperature influences the pH measurement through four
factors:
• Chemical equilibria are temperature dependent
• Temperature dependence of the slope (see Nernst equation
below)
• Position of the isothermal intersection
• Differing response time of the electrode (caused by temperature
changes).
pH Theory Guide
METTLER TOLEDO
35
pH Theory Guide
Every measuring solution has a characteristic temperature and pH behaviour (temperature coefficient). In general you should assume that a
temperature change results in a pH change (see buffer / temperature
table below). The reason for this is the temperature dependent dissociation which causes a change in the H+ concentration. This pH change
is real, not a measuring error.
Buffer / temperature table
Example:
20 °C
30 °C
0.001 mol / L HCI
pH 3.00
pH 3.00
0.001 mol / L NaOH
pH 11.17
pH 10.83
phosphate buffer
pH 7.43
pH 7.40
Tris buffer
pH 7.84
pH 7.56
This has to be taken into consideration if pH values obtained by different temperatures are to be compared.
The most exact pH value is obtained when the temperature of the
calibration and measured solutions are identical.
Furthermore, the slope depends on the temperature (Nernst equation):
E = E0 – 2.303 RT / F ∆ pH (R,F = constant)
Where: E = measured potential
E0 = zero point potential
∆pH = difference in pH between outside and
inside of the glass membrane
F = Faraday constant
R = universal gas constant
T = temperature Kelvin
the slope increases with an increase in temperature (see “5.6 The influence of temperature on pH measurements“ on page 85).
36
pH Theory Guide
METTLER TOLEDO
The temperature compensation of the transmitter corrects for this effect.
An electrode would have an ideal temperature behavior if its calibration
lines (isothermals) intersect at the zero point of the electrode
(pH 7 = O mV) at different temperatures (see Figure 18 below).
+mV
Real isothermal
intersection point
Theoretical isothermal
intersection point
Eis
0
pH
7
14
Error
T1
}
–mV
T2
T2 >T1
Figure 18 Calibration line and isothermal intersection points.
Since the overall potential of the pH electrode is composed of the sum
of many single potentials, which all have their respective temperature
dependencies, the isothermal intersection hardly ever coincides with
the zero point of the electrode (the ideal case would be: 0 mV at pH =
7 / 25 °C).
In the last few years, the development of the electrode has concentrated on bringing the isothermal intersection and the zero point as
close together as possible, since the nearer they are to pH 7 the
smaller the error in the temperature compensation. Besides, the measuring error increases with an increasing temperature difference be-
pH Theory Guide
METTLER TOLEDO
37
pH Theory Guide
tween the calibration and the actuai measuring. As a rule the errors are
in the order of 0.1 pH units (see “5.6 The influence of temperature on
pH measurements“ on page 85).
Temperature / time behavior of combination electrodes
When the temperature change of the medium is rapid, a conventional
pH electrode will drift until the temperature of the electrode and the medium become equal. In order for a combination electrode to react rapidly to the temperature changes of the medium, the temperature of the
inner lead-off element and the outer reference element must always be
identical. Alternatively the temperature dependance of the lead-off elements have to be equal to zero.
80°C
30°C
0.5
pHunits
Equit hal®
Convent ional
0
10
20
t (min)
30
Conventional electrode
Equithal® electrode
Asymmetrical lead-off
system of a conventional
electrode
Symmetrical lead-off
system of a Equithal®
electrode
Figure 19 Symmetrical structure of an Equithal®-system in comparison with a
conventional electrode.
38
pH Theory Guide
METTLER TOLEDO
Optimal electrodes are above all distinguished by the symmetrical
warming up or cooling down of their lead-off elements. They have the
same temperature coefficient and isothermal Intersection at pH 7 and
0 mV.
Thereby a short response time to temperature changes, as well as an
accurate temperature compensation can be guaranteed (see Figure 19
on page 38).
2.4
Troubleshooting
The troubleshooting diagram on page 40 should help the user to find and
eliminate an error or at least restrict the possible causes.
In the diagram the most frequent errors were considered. The diagram
is designed for troubleshooting industrial measuring systems.
2.4.1
Instructions and comments for the troubleshooting diagram
As a starting point for this diagram, the following situation has been
chosen:
• The on-line measurement differs from a comparison measurement in
the lab.
• After that, the electrode used in the on-line measurement has to be
calibrated (see “2.2.4 Calibration“ on page 30 and “5.5
Calibration / adjustment of the pH measurement setup“ on page 84).
The beginning of the troubleshooting diagram is «start».
Important: be sure to use correct buffer solutions to calibrate (see
chapters “2.2.4 Calibration“ on page 30 and “5.3 Buffer solutions“
on page 75).
pH Theory Guide
METTLER TOLEDO
39
Troubleshooting diagram
START
Path 1
Path 2
1
Y
Calibration
possible?
Measuring signal
virtually unchanged
in all solutions?
N
Temperature
compensation OK?
N
9
8
Y
2
Y
N
Short circuit
Slope too flat?
N
Y
12
3
Y
Stirring dependent
measuring signal?
Adjust
temperature
compensation
N
Y
N
N
Zero point
shift >30 mV?
11
Put short circuit
at amplifier
output
Follow path 1
N
Y
Replace
electrolyte
Immerse
electrode
in solution
Y Equal and stable
Crack in glass
electrode
Reading at its
limit?
Y
1
Y
Ground the
solution
Contaminated
junction
Slope too steep?
Y
Dry electrode
between junction
and membrane
Change of
signal?
10
N
Signal OK?
N
N*
5
Calibration OK?
value as before?
N
Y
1
4
Y
Clean junction
or replace
electrode
N
Change of
signal?
Put short circuit
at amplifier
output
Cable or plug
defect
Replace
electrode
Check
ground loop
Electrode o.k.
Y
Calibration OK?
N
6
Grounding
problem
Separate cable
from electrode
Poor isothermal
intersection
Replace
electrode
7
Choose
appropriate
electrode
Y
Change of
signal?
Cable defect
N
Y
Diagnosis
Action
Signal OK?
N
Replace
electrode
Separate cable
from transmitter
Change of
signal?
N
Check
transmitter
Instructions for use:
Question
Y
Check output
circuit for breaks
Check
transmitter
* Only valid for analog devices,
which have their monitor
connected to the output circuit.
Otherwise follow: N
Figure 20 Troubleshooting diagram
40
pH Theory Guide
METTLER TOLEDO
pH Theory Guide
METTLER TOLEDO
41
pH Theory Guide
42
pH Theory Guide
METTLER TOLEDO
When an error has been removed, go back to <> again in order
to find further possible errors. The following explanations have been
numbered according to the numbers in the boxes of the diagram:
1 The question “calibration possible” can be answered with yes
if the pH monitor allows the necessary corrections of the zero
point of the electrode and the slope. Many devices must have
a slope of the electrode of at least 85 % and at most 102 %
The zero point should not be more than ± 30 mV from the theoretical value (DIN standard). Only under these conditions
can the calibration be done (see “2.2.4 Calibration“ on page 30
and “5.5 Calibration / adjustment of the pH measurement
setup“ on page 84).
2 Temperature compensation OK: is the temperature of the
pH monitor correct, the temperature of the buffer
solution c orrectly adjusted? Are the temperature sensors,
when using automatic temperature compensation, in the
same solution (buffer) as the electrode?
(See “2.3.3 Temperature compensation“ on page 35 and “5.3
Buffer solutions“ on page 75)
3 A signal is said to be stirring-dependent if it changes when the
stirring velocity changes (or when the velocity of the medium
passing through a pipe line changes). Please notice that low
ionic strength measuring solutions may show a stirringdependent measuring signal independent of junction contamination.
4 See “2.3.1 Maintenance of the electrode function“ on page 32.
5 See “2.2.3 Signal processing and environmental influences“
on page 26.
6 An incorrect isothermal intersection may cause measuring
errors if the measuring temperature differs strongly from the
calibration temperature (see “2.3.3 Temperature compensation“ on page 35 and “5.6 The influence of temperature on pH
measurements“ on page 85).
7 Contact the electrode manufacturer
8 ”Slope too flat” means that the slope cannot be adjusted with
the applied device (when the slope is 85 % or less of the
theoretical value).
9 “Slope too steep” means that the slope cannot be adjusted
with the applied device (when the slope is 102 % or more of
the theoretical value).
10 Some devices allow a zero point correction of more than
± 30 mV (± 0.5 pH; 25 °C). According to DIN standards,
however, the zero point should not differ from the theoretical
value by more than ± 30 mV.
11 Please note: Only electrolytes recommended and supplied by
the electrode manufacturer for the corresponding electrodes
should be used to refill the electrode.
12 An artificial interruption (open circuit) is produced between the
glass and reference electrode.
pH Theory Guide
METTLER TOLEDO
43
pH Theory Guide
3
Intelligent Sensor Management
The most significant development in recent years in pH measurement
has been the introduction of digital measurement systems. These allow
a number of sensor installation, measurement, calibration, and maintenance advantages that analog systems cannot provide. METTLER
TOLEDO’s Intelligent Sensor Management (ISM) platform is one such
technology. ISM simplifies sensor handling, enhances process reliability, and reduces sensor lifecycle costs. Central to ISM is the inclusion
of a microprocessor in the sensor head. It is this feature that permits
ISM’s many benefits.
3.1
Signal integrity
As mentioned earlier in this guide, pH sensors transmit a high im
pedance mV signal to a transmitter which converts the signal to a
displayed pH level. Environmental influences such as humidity and
electromagnetic interference from surrounding equipment, plus the
length of cable from sensor to transmitter, can negatively impact the
integrity of the signal. In ISM sensors, the pH level is calculated in
the sensor’s microprocessor. Due to the proximity of the pH and reference electrodes to the microprocessor, the measured pH level is more
accurate than in analog systems. The digitized signal is then trans
mitted over coaxial cable to the transmitter for display and / or forwarding to a PLC system. Being digital, the signal is unaffected by environment and cable effects; therefore, the signal integrity of ISM systems
is extremely high. This makes ISM systems particularly suitable for
processes where the sensor and transmitter are far apart, and processes where there is a lot of moisture present in the environment.
44
pH Theory Guide
METTLER TOLEDO
3.2
Pre-calibration
A further advantage of ISM is the storing of sensor calibration data on
the sensor’s microprocessor. This means that sensors can be calibrated, using computer software (see iSense Asset Suite below) or an
ISM-equipped transmitter, in a convenient location and then stored until
they are required. This feature is particularly useful, for example, in the
biopharmaceutical industry where taking calibration fluids into a cleanroom presents a contamination risk, or in chemical processes where
there exists a potential health risk to employees.
Further, when connected to the transmitter, due to the calibration and
other configuration data held on the sensor, the new probe is instantly
recognized and the system is ready to measure in under a minute.
This Plug and Measure feature not only significantly reduces the time
for which a measurement system cannot be used, it eliminates the risk
of incorrect calibration data being entered into transmitters, therefore
increasing process reliability.
3.3
Predictive diagnostics
During batch processes, the failure of a measuring sensor can be
hugely detrimental. Ideally, operators want to know that a sensor will
operate correctly until the batch is complete. ISM technology is able to
provide expert diagnostic information that is particular to each measurement point. This information is available to operators via the display on ISM-equipped transmitters, or can be sent to the PLC.
The diagnostic tools allow a measurement point to be optimized on an
ongoing basis and all critical situations to be predicted so that operators can respond before production is interrupted.
pH Theory Guide
METTLER TOLEDO
45
pH Theory Guide
DLI
Dynamic Lifetime Indicator
TTM
DLI
Dynamic Lifetime Indicator
The DLI estimates in real time the remaining lifetime of the sensor. A unique algorithm uses current
and historic measurement and calibration values
Adaptive Calibration Timer
to calculate
the remaining lifetime of pH electrodes.
ACT
Dynamic Lifetime
Indicator
TTM
Time to Maintenance
ACT
Based on the DLI, the ACT calculates when the
next electrode calibration will be required.
Adaptive Calibration Timer
Adaptive Calibration
Timer
DLI
Dynamic Lifetime Indicator
Time to Maintenance
TTM
ACT
Adaptive Calibration Timer
The electrode also indicates when the next maintenance should be performed.
Time to Maintenance
Time to Maintenance
00/NN
CIP
CIP
CAL
CAL
CIP/SIP Counter
00/NN
CIP / SIP Counter
CIP/SIP Counter
Calibration History
CIP and SIP cycles are interpreted with a proprieMax. °C
Days of operation
tary, patented
algorithm.
MAX
MAX
Max.
Max.Temperature/ODI
°C
Days of operation
Max. Temperature/ODI
Calibration history is stored in the electrode and
can be used for diagnostics.
Calibration History
Calibration History
00/NN
CIP
CIP/SIP Counter
Max. °C
Days of operation
MAX
Max. Temperature/ODI
Maximum Temperature
46
CAL
pH Theory Guide
METTLER TOLEDO
Calibration History
Maximum Temperature / Operating Days Indicator.
Information about the maximum temperature the
sensor has ever been exposed to and the number
of operating days.
3.4
Asset management software
iSense Asset Suite is computer software for use with ISM sensors.
iSense provides a number of functions including sensor calibration,
electronic documentation, and management of all the ISM sensors at a
facility. iSense runs on PCs and laptops and is designed to be used
where sensor maintenance is most convenient. In chemical industry
applications this could be in a workshop or maintenance room which
is close to the production environment. In the pharmaceutical industry,
a laboratory room would be very suitable.
Calibration
Calibrating ISM pH electrodes using iSense is a straightforward and
quick process. Once completed, the calibration data is automatically
recorded in the history for the particular electrode. Calibrated probes
can then be stored until they are needed. Now when pH electrode
replacement is necessary it is a simple matter of taking a pre-calibrated probe from stock and a quick exchange at the measuring point.
No further calibration is required. There are no buffers at the process
and no risk of process contamination. iSense also contains a buffer
database which can be added to if required.
In addition, the digital nature of ISM allows for a very accurate calibration to be performed, which results in more reliable measurements and
excellent repeatability.
3.4.1
Electronic documentation
For the pharmaceutical and biopharmaceutical industries, meeting
stringent regulations includes being able to supply accurate records on
the calibration history of process analytical sensors used at the facility.
The time taken for technicians to transfer written notes to a PC can be
considerable, and there is always the possibility of human error in the
data transfer.
pH Theory Guide
METTLER TOLEDO
47
pH Theory Guide
With iSense, a record of an electrode’s calibration and maintenance
history is automatically stored in the internal database. Additionally,
data on maximum temperature exposure, operating time, number of
CIP / SIP cycles, etc. are also uploaded from the sensor to iSense. All
this data can be documented electronically or printed from a PDF.
A user management and electronic logbook allows control and tracking
of all activities ensuring a complete record of pH electrode assets over
their lifetime.
3.4.2
Sensor management
Every time an ISM pH electrode is connected to iSense, the software’s
Key Performance Table tells you if the electrode is “healthy”. The data
shown is appropriate for the electrode type and provides the most essential information, including sensor condition and history (e.g. exposure to CIP / SIP cycles). A smiley is shown on the screen to provide an
at-a-glance indicator of sensor state.
The key to effective sensor management with iSense, is the Dynamic
Lifetime Indicator (DLI) mentioned above. The unique algorithm behind
the DLI distills METTLER TOLEDO’s many years of experience regarding
the influence of process environment and maintenance (e.g. temperature, pH level, calibrations) on electrode lifetime and compares this historical data with the current process conditions. From this information
the DLI produces a real-time display of remaining electrode life. By
monitoring the condition of ISM assets you are able to ensure you always have sufficient stock of healthy sensors.
When the DLI indicates that an electrode can no longer be used, for
process security iSense can disable the sensor and issue a deactivation report. If a deactivated electrode is accidently connected at a
measuring point, the transmitter will recognize that the probe has been
disabled and it will not be accepted.
48
pH Theory Guide
METTLER TOLEDO
A complete view of all electrode data is provided by an SQL database
allowing you to access all previously stored data of all ISM sensors
used at your facility. Unlimited data export opens new possibilities for
optimizing measurement systems and processes, as well as maintenance management of the installed sensor base.
pH Theory Guide
METTLER TOLEDO
49
pH Theory Guide
4
Electrode selection and handling
For optimal pH measurements, the correct electrode must first be selected. The most important sample criteria to be considered are: chemical composition, homogeneity, temperature, process pressure, pH
range and container size (length and width restrictions). The choice becomes particularly important for non-aqueous, low conductivity, protein-rich and viscous samples where general purpose glass electrodes
are subject to various sources of error.
The response time and accuracy of an electrode is dependent on a
number of factors. Measurements at extreme pH values and temperatures, or low conductivity may take longer than those of aqueous solutions at room temperature with a neutral pH.
The significance of the different types of samples is explained below by
taking the different electrode characteristics as a starting point. Again,
mainly combination pH electrodes are discussed in this chapter.
4.1
Different kinds of junction
4.1.1
Ceramic junctions
The opening that the reference part of a pH electrode contains to maintain the contact with the sample can have several different forms.
These forms have evolved through time because of the different demands put on the electrodes when measuring diverse samples. The
“standard” junction is the simplest one and is known as a ceramic
junction. It consists of a porous piece of ceramic which is pushed
through the glass shaft of the electrode. This porous ceramic material
then allows the electrolyte to slowly flow out of the electrode, but stops
it from streaming out freely.
This kind of junction is very suitable for standard measurements in
aqueous solutions; the METTLER TOLEDO InPro 325x series is an example of such an electrode. A schematic drawing of the principle of
this junction is shown in Figure 21 below.
50
pH Theory Guide
METTLER TOLEDO
Figure 21 Electrode with ceramic junction.
Even though this is probably the most widely used junction because of
its simplicity of use with aqueous solutions, it has one main drawback:
Because of the porous structure of the junction it is relatively easy for
samples to block the junction, especially if the sample is viscous or if it
is a suspension.
You also have to be careful with some aqueous samples such as those
with a high protein concentration, since proteins may precipitate within
the porous junction if they come in contact with the reference electrolyte, which is often KCl. This reaction will cause the porous structure to
be filled with protein debris blocking the junction and rendering the
electrode useless. Measurements are not possible if the electrolyte cannot flow freely since the reference potential will no longer be stable.
The same problem can also be caused if the inner electrolyte reacts
with the sample solution being measured and the two meet in the junction. This reaction can create a precipitate which may block the junction, for example if KCl electrolyte saturated with AgCl is used with
samples containing sulfides, the silver and sulfides react to form Ag2S
which then blocks the ceramic junction. Factory-filled, prepressurized
liquid / gel electrolyte pH electrodes are suited to a wide scope of applications in the biotechnology, pharmaceutical and chemical process
industries. This ensures the best possible measurement performance
under the most diverse operating conditions.
pH Theory Guide
METTLER TOLEDO
51
pH Theory Guide
4.1.2
PTFE annular diaphragm
An annular PTFE diaphragm instead of a ceramic pot increase the surface to the media to prevent clogging on the diaphragm. PTFE is a dirtrepelling material.
Highly contaminated process conditions makes pH measurement and
control a complicated issue. An annular PTFE reference diaphragm
(e.g. METTLER TOLEDO‘s InPro 480x series) is designed for service in
tough environments. It resists fouling from hydrocarbon contaminants
and sulfides, ensuring high accuracy and fast response throughout its
long life. For process media containing particles and aggressive chemicals, the optional flat glass membrane electrode is the optimal
solution.
PTFE
diaphragm
pH
membrane
glass
Measuring
solution
long diffusion path
Figure 22 Example of electrode with PTFE diaphragm.
52
pH Theory Guide
METTLER TOLEDO
Open junctions
The third type of junction is the open junction. This reference electrode
is completely open to the environment and has full contact between the
reference electrolyte and the sample solution. This is only possible with
a solid polymer reference electrolyte. A schematic diagram of this junction is shown below.
Glass
4.1.3
Open junction
Measuring
solution
Xerolyt® Extra
solid polymer
electrolyte
Figure 23 Example of electrode with open junction.
The great advantage of this junction is clearly the fact that it is completely open and therefore is unlikely to clog. Open junctions can easily
cope with very dirty samples constantly providing good measurements. The disadvantage of the solid polymer reference electrolyte
which is used for this open junction is that it has slower reaction times
and low electrolyte flow. This means that the samples measured need
to have a high enough ion concentration for stable measurements to
be possible. Nevertheless, these electrodes are suitable for most samples and are very robust.
4.1.4
Dual-membrane without junction
The cell membrane chlor-alkali process is very tough on conventional
pH electrodes. It exposes them to high temperatures, and clogging and
poisoning from a variety of compounds. This is particularly true in the
anode side of the electrolysis cell. Chlorine diffuses through the electrode’s diaphragm and attacks the reference system. This results in incorrect pH measurement and shorter sensor lifetime.
pH Theory Guide
METTLER TOLEDO
53
pH Theory Guide
pH-sensitive glass
Sodium-sensitive glass
Figure 24 Dual-membrane pH electrode.
Reliable pH measurement can be achieved with sensors such as the
InPro 4850 i from METTLER TOLEDO. This is a dual-membrane pH
electrode that has been designed to provide long-term accurate measurement in chlor-alkali processes. The main difference in measuring
technology between dual-membrane pH electrodes and conventional
pH electrodes is the presence of a sodium-reference (pNa) system.
The electrode features a sodium-sensitive glass membrane which is
charged by the sodium ions in the process medium. The sodium concentration in the brine is used as a reference. The pNa reference system
is hermetically sealed; there is no diaphragm, therefore no oxidants
can enter the electrode and attack the reference system. The electrode
also features a high-alkali resistant pH membrane glass for pH measurement. It is the amalgamation of pH measurement and pNa reference that is one reason that this kind of electrode is highly suited to
chlor-alkali processes.
4.2
Reference systems and electrolytes
Of all the possible reference systems developed for reference elements,
only a few are of practical importance. These are the silver / silver
chloride, iodine / iodide and the mercury / calomel systems, as well as
some of their adaptations. Due to environmental considerations,
however, the calomel reference electrode is no longer widely used.
Here we only discuss the most important reference system, the silver / silver chloride system.
54
pH Theory Guide
METTLER TOLEDO
The potential of the reference electrode system is defined by the reference electrolyte and the reference element (silver / silver chloride). The
conventional construction of this reference system is a silver wire
coated with AgCl. For this version of the Ag / AgCl reference system it is
important that the reference electrolyte has a very high (saturated) AgCl
concentration to ensure that the reference element wire does not get
stripped of the AgCl. If this were to happen the reference element would
stop working.
An improvement to this type of reference element was made with the development of the ARGENTHAL™ reference element. The ARGENTHAL™
reference element consists of a small cartridge filled with AgCl particles
that provide the silver ions for the chemical reaction at the lead off wire.
This cartridge contains enough AgCl to last the lifetime of the electrode.
Which type of reference electrolyte is used in an electrode strongly depends on the reference system and on the type of sample used.
Silver wire coated with AgCl
Ag/AgCl cartridge (ARGENTHAL™)
Glass wool
Silver ion trap
Ag+ free reference electrolyte
Diaphragm
Figure 25 Schematic drawing of the ARGENTHAL™ reference system.
Whereas the reference system can either be conventional silver wire or
ARGENTHAL™, the sample can be divided into two classes namely
aqueous and non-aqueous matrices.
For both aqueous and non-aqueous solutions it is important that the
reference electrolyte contain plenty of ions to keep the reference system
working well. Ideally, the salts used to provide these ions in the referpH Theory Guide
METTLER TOLEDO
55
pH Theory Guide
ence electrolyte are very soluble in the solvent, are pH neutral (so that
they do not influence the measurements when flowing out of the electrode) and do not precipitate out by reacting with other ions present in
the sample or buffer. KCl matches these requirements best for aqueous
solutions and LiCl is best suited for use with non-aqueous solutions.
The conventional Ag / AgCl reference system needs the presence of an
electrolyte saturated with AgCl (see Figure 25 on page 55) so that the
lead off wire does not get stripped of AgCl. The reference electrolyte of
choice is therefore, 3 mol / L KCl saturated with AgCl. The disadvantage
of this electrolyte is that silver ions can react with the sample to form
an insoluble precipitate thereby blocking the junction.
As mentioned previously, the ARGENTHAL™ reference system has a
cartridge with AgCl granules which ensure that AgCl is constantly available. Typically, this ARGENTHAL™ system comes in combination with a
silver-ion barrier which stops silver ions from passing into the electrolyte. The advantage of these features of the ARGENTHAL™ reference
system is that you can use standard 3 mol / L KCl as a reference electrolyte rather than 3 mol / L KCl saturated with AgCl, so in combination
with the silver-ion trap there are no free Ag+ ions in the electrolyte
which could cause a precipitate after reaction with the sample.
A phase separation in the contact area between electrolyte and sample
solution at the junction can cause an unstable signal, therefore deionized water is used as a solvent for the reference electrolyte in aqueous
samples, and ethanol or acetic acid is used as solvent for non-aqueous systems.
56
pH Theory Guide
METTLER TOLEDO
A brief overview of the possible reference system / electrolyte combinations is given below:
Electrolyte for aqueous samples
Electrolyte for nonaqueous samples
ARGENTHAL™
Conventional
ARGENTHAL™
3 mol / L KCl + H2O
3 mol / L KCl + AgCl + H2O
LiCl + Ethanol /
LiCl + Acetic acid
In addition to the above-mentioned liquid electrolytes, there are also
gel and solid polymer electrolytes. Electrodes delivered with these electrolytes cannot be refilled.
The electrode response time is strongly dependent on the type of electrolyte used. Liquid electrolyte electrodes show a very quick response
time and give the most accurate measurements. Gel and solid polymer
electrolyte electrode both have longer response times, but they are virtually maintenance-free.
pH Theory Guide
METTLER TOLEDO
57
pH Theory Guide
4.3
Types of membrane glass and membrane shape
The pH glass membrane of an electrode can have several different
shapes and properties, depending on the application the electrode is
used for. The selection criteria here are sample consistency, volume
and temperature, the required measurement range and the concentration of ions present in the sample.
The most obvious property is the shape of the membrane, and in
Figure 26 below a selection of membrane shapes is shown together
with their properties and proposed usage.
Spherical
Hemispherical
For low temperature
samples: resistant to
contraction
Small sample
volume: pH membrane
only on the bottom
Spear
Flat
For semi-solid and solids:
punctures the sample
easily
For surfaces and drop
sized samples: very
small pH-membrane
contact area
Cylindrical
Highly sensitive
membrane: large
surface area, lower
resistance
Micro
Samples in reaction
tube: very narrow
electrode shaft
Figure 26 Differently shaped pH membranes.
The membrane glass is also important for the measurement properties
of the electrode. The table below gives an overview of the various types
of METTLER TOLEDO pH membrane glasses.
58
pH Theory Guide
METTLER TOLEDO
Type of membrane glass
Properties / samples
HA:
High alkali glass
For high temperatures and high pH
values: extremely low alkali error
LoT:
Low temperature glass
For low temperatures and low ion
concentrations: low resistance glass
A41
For high temperatures; resistant to
chemicals
HF:
Hydrofluoric acid-resistant glass
For samples containing hydrofluoric
acid (up to 1g / L)
Na:
Sodium-sensitive glass
Only used for sodium detecting
electrodes: sodium specific glass
The HF membrane glass electrode is more robust in solutions with
hydrofluoric acid than standard pH electrodes. Hydrofluoric acid above
certain concentrations (> 1g / L) and below pH 5 attacks glass and prevents the development of a gel layer on the standard pH glass membrane. This again leads to unstable measurement values and also reduces the life span of the electrode.
4.4
pH electrodes for specific applications
Now that we have seen what different types of junctions, electrolytes
and membranes exist in pH electrodes, we will have a look at what this
means for the measurement of the pH in different systems.
4.4.1
Highly accurate problem solver
A highly accurate problem solver pH electrode is sufficient for routine
measurements in most applications where a lot of aqueous solutions
are tested. The advantage of this kind of pH electrode is that it is very
easy to use and is also very robust. In general, these electrodes are
made of glass and have a ceramic junction. They are also refillable,
which means that you can refill the electrolyte thereby cleaning the
electrode and prolonging its lifetime. For example, an electrode of
choice for these simple measurements is the InPro 200x (i).
pH Theory Guide
METTLER TOLEDO
59
pH Theory Guide
Figure 27 InPro 200x (i)
4.4.2
Complex samples or such of unknown composition
Measuring the pH of complex samples can be somewhat tricky, since
dirt in the sample can hinder correct measurements. Examples of such
applications are soil acidity measurements, quality control in foodstuffs
such as soups and measurements in colloidal chemical systems. The
risk of blockages with such samples would be very high if you were to
use a pH electrode with a ceramic junction. Therefore it is best to use a
pH electrode with an open junction such as the InPro 426x which has
a solid state polymer reference electrolyte Xerolyt® Extra (Mettler
Toledo patent). This electrode has a hole in the shaft which allows
direct contact between the electrolyte and sample.
Figure 28 InPro 426x (i)
60
pH Theory Guide
METTLER TOLEDO
It offers the following advantages:
• Insensitivity to flow variations.
• Especially adapted for low ionic content media and organic solvents.
• Particularly insensitive to anionic salt contents such as nitrate, sulfate, carbonate.
• Dual-open junction between sample medium and reference electrolyte (no diaphragm means no clogging) allows reliable measurements in heavily contaminated process media, in suspensions and
emulsions, and in solutions containing protein.
• Particularly efficient in sulfide-bearing media.
4.4.3
Semi-solid or solid samples
Standard pH electrodes are generally not able to withstand the pressure
of being pushed into a solid sample; therefore you need a special electrode which is able to penetrate the sample in order to measure the pH.
The shape of the membrane is also important as it needs to be formed
in such a way as to ensure a large contact area with the sample, even
if the electrode is pushed into the sample with force.
The METTLER TOLEDO electrodes most suitable for these kinds of applications are the Puncture pH Electrodes. While their spear-shaped
point enables them to pierce the sample, the membrane shape ensures
accurate measurements. The Puncture pH Electrode also has an open
junction, which further prevents the junction from being blocked by the
(semi-) solid sample. This electrode is typically used for quality control
or checking production processes of cheese and meat.
Figure 29 Puncture pH electrode
4.4.4
At the toughest applications in chemical p rocess industries
With a long diffusion path, pH electrodes can be highly resistant to oxidizing media, solvents, and acid or alkali solutions, and they enable
operation at particularly high process pressures. A dirt repellent annular junction (PTFE) can prevent the dirt clogging on the diaphragm
surface.
pH Theory Guide
METTLER TOLEDO
61
pH Theory Guide
METTLER TOLEDO‘s InPro 480x offer reliable measurements in oxidizing media, in strong acid or alkali solutions, also at high process
pressures and temperatures. The flat glass membrane version with an
integrated auxiliary platinum electrode (SG) is available for media containing a high amount of particles.
Figure 30 InPro 480x (i)
4.4.5
Prepressurized electrolyte pH electrodes
These electrodes have been designed for use at medium operating
pressures and in particular where reliable and highly accurate measurements are of special importance.
METTLER TOLEDO‘s InPro 325x (i) series is available with an expanded selection of different pH-sensitive glass membranes. This series delivers high measurement performance under the most diverse
operating conditions, both in chemical and biotech processes. The
following features make the InPro 325x series of pH electrodes dependable measurement tools in demanding applications:
• Extended operating life and precise measurement values:
Any ingress of process medium into the reference system is avoided
by (pre-)pressurization of the liquid electrolyte. Permanent overpressure within the electrode ensures that the diaphragm is continuously
cleaned by the action of the constant outflow of small amounts of
electrolyte through the diaphragm.
62
pH Theory Guide
METTLER TOLEDO
• Resistance to potential measurement problems at the diaphragm:
The silver-ion barrier integrated in the reference system prevents contamination or plugging of the diaphragm by black silver sulfide during measurements in process media containing sulfides or amino
acids.
Figure 31 InPro 325x (i)
4.4.6
Dual-membrane pH electrodes
The chlor-alkali process is very tough on conventional pH electrodes.
It exposes them to high temperatures, and clogging and poisoning
from a variety of compounds. Chlorine diffuses through the electrode’s
diaphragm and attacks the reference system. This results in incorrect
pH measurement and shorter sensor lifetime. METTLER TOLEDO’s
InPro 4850 i is a combination pH electrode featuring a sodium membrane glass that uses the sodium concentration in the process (brine)
as a reference. The difference in electrical potential between the pH
glass and the sodium reference glass is calculated into the pH value.
The sodium reference system is highly resistant to chlorine and other
oxidizing agents. This makes the sensor very well suited for the demanding process conditions in chlor-alkali production. Analog to digital
signal conversion ensures 100 % signal integrity and stability.
pH Theory Guide
METTLER TOLEDO
63
pH Theory Guide
Figure 32 InPro 4850 i
4.4.7
pH measurements in high purity water samples
Pressurized gel-filled electrodes provide greater stability of the reference
diaphragm / junction potential by forcing a small amount of potassium
chloride gel through it. The METTLER TOLEDO Thornton pHure
Sensor™ system offers this type of electrode. It requires no maintenance other than occasional calibration throughout its one year life.
Figure 33 pHure Sensor™
64
pH Theory Guide
METTLER TOLEDO
4.4.8
Installation in an upside-down position
An air-cushion system in the reference electrolyte eliminates any possible disturbance at the diaphragm caused by air bubbles in the electrolyte. METTLER TOLEDO’s InPro 3100 UD is intended for bottom-entry
installation.
Figure 34 InPro 3100 (i)
4.4.9
Non-Glass (ISFET) pH electrodes
Unbreakable solid-state ISFET pH electrode specially intended for use in
the food processing industry, where the possibility and consequences
of fractured glass present a threat. METTLER TOLEDO’s InPro 3300 is
equipped with a solid-state pH sensitive part (ISFET) and with a high
temperature reference system with gel electrolyte, making the electrode
practically maintenance-free. It is pressure-resistant up to 6 bar at
130 °C (PED 97) and fully sterilizable, either in-situ or in an autoclave.
The 45° design of the electrode end prevents air bubbles being trapped
at the pH-sensitive tip.
pH Theory Guide
METTLER TOLEDO
65
pH Theory Guide
Figure 35 InPro 3300 (ISFET pH sensor)
4.4.10 For low maintenance and simple installation
Polymer-body electrodes such as METTLER TOLEDO‘s InPro 4501/ 4550
are especially designed for high process pressures and temperatures,
with a special resistant polymer shaft material (PVDF or PPS) and
high-temperature glass quality. NPT threads allow the sensor to be
screwed directly into immersion tubes, pipes, process vessels, etc.
Figure 36 InPro 4501
66
pH Theory Guide
METTLER TOLEDO
Figure 37 InPro 4550
4.5
Electrode maintenance
Regular maintenance is very important for prolonging the lifetime of
any pH electrode. Electrodes with liquid electrolyte need the electrolyte
to be topped-up when the level threatens to become lower than the
level of the sample solution. This maintenance prevents a reflux of the
sample into the electrode. The complete reference electrolyte should
also be changed regularly, for example once a month. This ensures
that the electrolyte is fresh and that no crystallization occurs despite
evaporation from the open filling port during measurement.
It is important not to get any bubbles on the inside of the electrode,
especially near the junction. If this happens the measurements will be
unstable. To get rid of any bubbles, gently shake the electrode in a
vertical motion as with a fever thermometer.
4.6
Electrode storage
Electrodes should always be stored in aqueous and ion-rich solutions.
This ensures that the pH-sensitive gel layer which forms on the pH
glass membrane remains hydrated and ion rich. This is necessary for
the pH membrane to react in a reliable way with respect to the pH
value of a sample.
pH Theory Guide
METTLER TOLEDO
67
pH Theory Guide
4.6.1
Short term storage
In between measurements or when the electrode is not being used for
brief periods of time, it is best to keep the electrode in a holder containing its inner electrolyte solution (e.g. 3 mol / L KCl), or in a pH 4 or
pH 7 buffer. Ensure that the level of solution in the beaker is below that
of the filling solution in the electrode.
4.6.2
Long term storage
For long term storage, keep the electrode wetting cap filled with the inner electrolyte solution, pH buffer 4 or 0.1 mol / L HCl. Make sure that
the filling port for reference and combination electrodes is closed so as
to avoid loss of the electrolyte solution through evaporation, which can
cause the formation of crystals within the electrode and junction.
Never store the electrode dry or in distilled water as this will affect the
pH-sensitive glass membrane and thus shorten the lifetime of the
electrode.
Although regeneration can restore an electrode that has been incorrectly stored, following these recommendations will ensure that your
electrode is always ready to use.
4.7
Electrode c leaning
To clean the electrode, rinse it with deionized water after each measurement but never wipe it clean with a tissue. The surface of the paper
tissue can scratch and damage the pH-sensitive glass membrane, removing the gel-layer and creating an electrostatic charge on the electrode. This electrostatic charge causes the measured signal to become
very unstable. Special cleaning procedures may be necessary after
contamination with certain samples. These are described in greater detail below.
68
pH Theory Guide
METTLER TOLEDO
4.7.1
Blockage with silver sulfide (Ag2S)
If the reference electrolyte contains silver ions and the sample being
measured contains sulfides, the junction will become contaminated
with a silver sulfide precipitate. To clear the junction of this contamination, clean it with 8 % thiourea in 0.1 mol / L HCl solution.
4.7.2
Blockage with silver chloride (AgCl)
The silver ions from the reference electrolyte can also react with samples that contain chloride ions, resulting in an AgCl precipitate. This
precipitate can be removed by soaking the electrode in a concentrated
ammonia solution.
4.7.3
Blockage with proteins
Junctions contaminated with proteins can often be cleaned by
immersing the electrode into a pepsin / HCI (5 % pepsin in 0.1 mol / L
HCl) solution for several hours.
4.7.4
Other junction blockages
If the junction is blocked with other contaminations, try cleaning the
electrode in an ultrasonic bath with water or a 0.1 mol / L HCl solution.
4.8
Electrode regeneration and lifetime
Even electrodes that have been well maintained and properly stored
may start performing poorly after some time. In such cases it may be
possible to regenerate the pH-sensitive glass membrane and restore
the electrode to its previous level of performance using an ammonium
bifluoride regeneration solution. This regeneration solution is based on
a highly diluted solution of hydrofluoric acid which etches away a very
thin layer of the glass membrane, exposing a fresh surface area.
When using the regeneration mixture do not leave the electrode in the
solution for longer than 1-2 minutes or the entire pH-sensitive membrane will be corroded away and the electrode rendered useless.
pH Theory Guide
METTLER TOLEDO
69
pH Theory Guide
70
pH Theory Guide
METTLER TOLEDO
The expected lifetime of a correctly used and maintained pH electrode
is around one to three years. Factors that contribute to a reduction of
the lifetime of an electrode include high temperatures and measuring at
extreme pH values.
5
Comprehensive pH theory
In the previous sections the practical aspects of pH measurements
were discussed. This chapter will principally deal with the theoretical
background to pH measurements and is intended for readers wishing
to acquire a more fundamental understanding of pH theory. First, the
basic pH theory is developed, then we will have a look at the sensor
theory and at the end some special topics will be dealt with.
5.1
Definition of the pH value
According to Sørenson the pH is defined as the negative logarithm of
the H3O+ ion activity:
pH = – log [aH+]
From the equation we can see that if the H3O+ ion activity changes tenfold, the pH value changes by one unit. This nicely illustrates how important it is to be able to measure even small changes in the pH value
of a sample.
Often, the pH theory is described with H+ ions in connection with pH
values, although the correct ion to refer to is the hydronium (or as it is
officially known according to IUPAC: oxonium) ion (H3O+):
H+ + H2O ↔ H3O+
Not only acids and bases show dissociation behavior to form hydronium ions or hydroxide ions, but pure water also dissociates to form
hydronium and hydroxide ions:
2 H2O ↔ H3O+ + OH–
pH Theory Guide
METTLER TOLEDO
71
pH Theory Guide
The dissociation constant for this behavior is called Kw and is also
known as the autoionization or autodissociation of water:
Kw =
[H3O+] [OH–]
= [H3O+] [OH–] = 10 –14 (25 ºC)
[H2O]
From the Kw equation we can see that when equal amounts of H3O+
and OH– are present the solution is neutral, and this is the case when
both [H3O+] and [OH–] are 10 –7 mol / L, so at pH 7. When a higher concentration of H3O+ ions is present, then the pH value goes into the
acidic region of the pH scale, for example a H3O+ concentration of
10 –3 mol / L (and thus [OH–] = 10 –11 mol / L) gives a pH value of 3.
To be able to measure this value in a sample solution we need to know
how pH sensors react to the acid concentration in the solution. We will
examine this later in this chapter.
5.2
Correlation of concentration and activity
Up to now we have only discussed the concentration of acids and
bases as the determining factor for the pH value measurement. In reality though what is actually measured by a pH sensor is the activity of
the hydronium ions in solution. The concentration is only used, as in
many other chemical processes, as a simplification for using the activity of a solution. In many conditions the use of the concentration is a
very good approximation to using the activity.
The activity of the hydrogen ion (aH+) is defined by the concentration of
hydrogen ions and the activity coefficient (γH+). The concentration in
this case is usually given as the molality (b = mol / kg solvent) and not
the molarity (c = mol / L solution), as molality is a less ambiguous definition. The activity is then given by:
aH + = γ H + b H +
In dilute solutions the approximation aH+ = bH+ can be made.
72
pH Theory Guide
METTLER TOLEDO
The activity coefficient is not a universal constant; the value of this
number again depends on various factors such as temperature (T), total ion strength (I), the dielectric constant, ion charge (z), the size of
the ions (in Angstroms) and also on the density (d) of the medium.
There are two main effects which can be observed when noting the difference between ion activity and ion concentration. These are the socalled salt effect and medium effect.
The influence of salts present in a solution of which the pH value is
measured is called the salt effect. This salt effect is denoted by the
symbol γxH+
and is defined as:
– 0.5 I½
log γxH+ =
1 + 3 I½
In this equation “I” is the symbol for the total ionic strength = ½ ∑ ci zi2
If we assume in the case of pH measurement that both the anion and
the hydrogen ion are monovalent, zi will be equal to 1 and the total ion
strength I is determined by the molality. The influence of the salt effect
on the activity coefficient of selected ion concentrations is shown in the
following table.
Molality
0.001
0.005
Activity
coefficient
0.01
0.967294
0.935044 0.915247
0.05
0.1
0.857205
0.829586
pH Theory Guide
METTLER TOLEDO
73
pH Theory Guide
When we now compare a pH measurement in a solution of 0.01 mol / L
HCl with or without salt present, we get the following comparison:
0.01 mol / L HCl solution
0.01 mol / L HCl solution
with 0.09 mol / L KCl
pH = – log (bH+ γxH+)
= – log (0.01 0.915)
= – log (9.15 10 –3)
= 2.04
pH = – log (bH+ γxH+)
= – log (0.01 0.829)
= – log (0.829 10 –3)
= 2.08
From this example it can be seen that the pH value increases by
0.04 pH units (the H+ activity decreases) in solutions with a higher ion
strength. This explains why solutions with the same acid content may
show different pH values if there are other salt ions present in the solution.
The second effect which links activity to concentration is the so-called
medium effect. The medium effect is designated with:
γmH+
This effect shows what influence the medium (solvent, etc.) will have
on the H+ ion activity. With this effect electrostatic and chemical interactions play an important role. For example, the H+ activity is 200
times greater in ethanol than in water.
When taking both the salt effect and medium effect into account, the
relationship between concentration and activity then becomes:
aH+ = γxH+ γmH+ bH+
From these examples we can see that it is very important to have detailed knowledge of the measured sample, since the more accurately
defined the measuring conditions are, the more reproducible the pH
values obtained will be.
74
pH Theory Guide
METTLER TOLEDO
5.3
Buffer solutions
Buffer solutions are a very important part of an accurate pH measurement. Standard buffers are used to calibrate pH sensors and to check
their performance. The most important property of a pH buffer is its
buffering capacity, hence its name. This property enables a pH buffer
to remain at a constant pH value, even if external substances are introduced into the buffer solution.
The buffering capacity of a buffer solution depends on the fact that
weak acids only partly dissociate, causing the following equilibrium reaction:
HA ↔ H+ + A –
In this equilibrium, the anion A – can act as a base, since it can withdraw protons from the system. The non-dissociated acid HA, however,
can supply the system with protons.
A buffer solution in its equilibrium state therefore has enough anions
(A –) to take up any protons added to the system, but also has enough
non-dissociated acid available to replace any protons withdrawn from
the system. Since the non-dissociated acid HA can act as an H+ donor
and the dissociated acid A – as an H+ acceptor, a buffer solution will be
at its most powerful when both HA and A – are present in equal concentrations.
If we first have a closer look at the theory of buffer solutions, we
can then find out how suitable a certain solution is as a buffer. This
depends on several properties of the buffer solution, such as buffer
capacity, temperature influences, and changes of the pH value due
to dilution of the buffer solution. These properties are documented for
many standard buffer solutions and can be found in the literature.
pH Theory Guide
METTLER TOLEDO
75
pH Theory Guide
From the formula above we can write the equilibrium constant for a dissociated acid as follows:
Ka =
[H+] [A –]
[HA]
This can be then be re-written as:
1 1 [A –]
=
[H+] Ka [HA]
and then taking the logarithm on both sides:
1
1
[A –]
log + = log
+ log
[H ]
Ka [HA]
Since
1
log
= – log [H+] = pH
[H+]
and
1
log
= – log Ka = pKa
Ka
we then get:
[A –]
pH = pKa + = log
[HA]
This equation is known as the HENDERSON-HASSELBALCH equation.
From this last equation we can see that if a buffer solution is at its
strongest and therefore [A –] = [HA], that the pH value corresponds to
the negative log of the dissociation constant,
pH = pKa
This equation is very helpful when making a buffer solution of a weak
acid with known pKa value.
76
pH Theory Guide
METTLER TOLEDO
5.3.1
Buffer capacity (ß)
The buffer capacity is defined as the ability of a buffer solution to maintain its pH value even after the addition of a strong acid or base.
As we have seen in the previous section, the greatest buffer capacity is
when pH = pKa, but the overall buffer capacity of a weak acid or base
is limited to pH = pKa ± 1.
As an example of the buffer capacity of a weak acid we will look at a
titration curve of acetic acid (CH3COOH) with OH– ions titrated into the
solution (Figure 38). Acetic acid has a pKa value of 4.8, so this solution starts with a low pH value and the pH value increases when more
hydroxide ions are titrated into the solution. At the beginning the
change is quite big with every drop of hydroxide solution, but when the
concentrations of the non-dissociated acid and dissociated acid start
becoming equal the curve becomes flatter. As [A –] = [HA] when pH =
pKa, we expect the curve to become flat around pH 4.8, since this is
the pH value where the buffering capacity should be most pronounced.
pH
▲
4.8
▲
[A–]/[HA]
Figure 38 Buffering capacity of acetic acid.
When making and using buffer solutions you have to be aware of external influences on the acid / base equilibrium as well. One example of
this could be the uptake of CO2 from the air.
pH Theory Guide
METTLER TOLEDO
77
pH Theory Guide
5.3.2
Dilution value (∆pH)
The dilution value of a buffer solution indicates how much the pH value
changes when the buffer solution is diluted with an equal amount of
distilled water.
A positive dilution value means that the pH will increase whereas a
negative dilution value means that the pH will decrease with increasing
solution.
5.3.3
Temperature effect (∆pH / ∆T)
We have seen the pH value is derived from the activity of the H+ ions in
the solution. Since the ion activity is temperature dependent, the temperature will also influence the pH value.
The temperature coefficient expresses changes of the pH value per °C.
5.4
The measurement chain in the pH measurement setup
We saw in chapter “1.3 The tools for pH measurements“ on page 14, that
a pH measurement is actually the measurement of a potential. The
changing potential of a pH-sensitive electrode is measured against the
stable potential of a reference electrode. A measurement setup is
shown in Figure 7 on page 15.
The principle of the setup is that metal conductors within the two electrodes are connected to each other through one or more electrolytes to
form a galvanic chain. To this galvanic chain (pH and reference electrode) a meter with a high input resistance is attached and this connects the two electrodes internally and measures the chain potential E.
This galvanic potential E is defined by the Nernst equation:
RT
log aH+
E = E0 + 2.3
nF
which we have seen before in Figure 6 on page 14.
78
pH Theory Guide
METTLER TOLEDO
In order to be able to compare the galvanic potentials of different electrodes with different reference systems, the standard hydrogen electrode (SHE) or normal hydrogen electrode (NHE) was introduced as a
universal reference electrode. The potential of the SHE is by definition
zero at all temperatures. The SHE consists of a platinized platinum
sheet which is immersed in a solution of aH+ = 1.0 and surrounded by
hydrogen gas at 1 bar.
In the Nernst equation E0 is the standard potential at aH+ = 1.
The factor 2.3 RT / nF (EN) is the slope of the pH electrode and gives
the change in measured potential with tenfold change in H+ activity, or
per pH unit. The value of EN depends on the temperature T in Kelvin,
and is often referred to as the slope factor. Some examples for the
slope at certain temperatures are given in Figure 39 below.
Temperature
EN Value (mV) / pH
0 °C
EN = 54.2 mV / pH
25 °C
EN = 59.2 mV / pH
50 °C
EN = 64.1 mV / pH
Figure 39 Temperature dependence for the pH electrode slope factor.
When we look at the measurable chain potential E from the Nernst
equation in a bit more detail, we find that this chain potential consists
of several intermediate potential points, which are shown in Figure 40
below.
E
E4
Reference
electrolyte
E3
E2
E5
E6
Inner
buffer
E1
Figure 40 Different sources of potential in a combination electrode.
pH Theory Guide
METTLER TOLEDO
79
pH Theory Guide
5.4.1
pH electrode
The chain potential starts at the contact area between the sample solution and the pH electrode glass membrane, where the potential E1 is
measured in correlation with the pH value of the sample solution. In order to measure E1 and assign a definite pH value to it, all other single
potentials in the chain E2-E6 have to be constant. The only variable
signal is caused by the potential difference between inner electrolyte
and sample solution over the pH membrane. The last point in the chain
is E6, the potential between the reference electrode electrolyte and the
sample solution again, which has a constant potential since the reference electrode is insensitive to the pH value of the sample.
The other potentials E2, E3, E4, and E5 are the consecutive steps in the
chain from the sample through the pH electrode to the meter, and back
again from the meter through the reference electrode to the sample solution. All these separate steps can be seen in Figure 40 on page 79.
The potential E1 is transferred to the inside of the pH membrane glass
via the gel layer on the glass membrane and the pH glass membrane
(as shown in “Figure 8 Cross sections through the glass membrane.“
on page 16), where another gel layer is present as an interface between
the inside of the pH electrode and the inner buffer solution. The potential difference between the outside of the pH glass membrane and the
inside of the pH glass membrane is the potential E2 in Figure 40 on
page 79.
Physically this works by transferring the potential via an equilibrium of
the hydrogen ions which arises at the interface between the measuring
solution and the outer pH membrane gel layer. If the activity of the hydrogen ions is different in the two phases, hydrogen ion transport will
occur. This leads to a charge at the phase layer, which prevents any
further H+ transport. This resulting potential is responsible for the different hydrogen ion activities in the sample solution and the gel layer. The
number of hydrogen ions present in the gel layer is given by the silicic
acid skeleton of the glass membrane and can be considered a constant and therefore independent of the measuring solution.
80
pH Theory Guide
METTLER TOLEDO
The potential in the outer gel layer of the pH-sensitive membrane is
then transferred by the Li+ ions found in the glass membrane to the inside of the glass membrane, where another phase boundary potential
arises (E3 in Figure 40 on page 79).
The potential E3 is then transferred to the lead-off wire in the pH electrode (E4) via the inner buffer solution of the pH electrode and from
there to the meter.
5.4.2
Reference electrode
When the pH electrode potential chain (E1-E4) signal goes to the
meter, there needs to be a reference signal available in the meter as
well to measure the pH signal against. This is done with the reference
part of the electrode, where another potential chain (E5-E6) ensures
this stable potential independent of the sample solution.
From the meter there is a connection to the reference element of the
reference electrode and from there an interface between the reference
element and the reference electrolyte solution (potential E5).
Of the different reference elements, the silver / silver-chloride element
has become the most important one. Compared to the calomel
electrode the silver / silver-chloride reference has some important advantages, but it is mainly because of environmental reasons that
the calomel reference electrode has almost completely disappeared.
The next step is the potential E6, which is the connection between the
reference electrolyte on the inside of the reference electrode and the
sample solution on the outside of the electrode. Again, it is important
that the potential is stable here as it is used as a reference signal. The
junction is naturally very important for this particular contact since it allows the diffusion of the ions through the junction.
The critical property of the junction is the diffusion of ions through it
which generates the diffusion potential (E6 / Ediff). The diffusion potential depends not only on the type of junction and its properties, but also
on the diffusing ions.
pH Theory Guide
METTLER TOLEDO
81
pH Theory Guide
Since Ediff is a part of the potential in every measuring chain, the pH
values of different measuring solutions can, strictly speaking, only be
compared if the diffusion potential is identical in all solutions. In practice this is not always possible, so it is important to keep Ediff small
and constant to limit the measurement error.
The migration velocity of ions is determined by their charge and size.
The size of an ion is determined not by its “net” size, but by the size of
its h ydration cover. All ions in aqueous solutions are surrounded by
polar water molecules. This means that a small but highly hydrated
lithium ion for example migrates slower than a much larger but only
slightly hydrated potassium ion. Since the H+ and the OH– ions migrate
in accordance with completely different mechanisms, they have a
much higher ion mobility compared to all other ions. Examples of migration speeds for different ions are shown in Figure 41 below.
Ionic mobilities (in 10 –4 cm2 / s · V) at 25 °C
H+
Li
36.25
+
Na
4.01
+
K+
NH4
+
OH–
F
5.74
–
5.19
Cl
7.62
NO3–2
7.62
20.64
–
CH3COO
7.91
7.41
–
4.24
Junction
▲
Na+
Solution 2
Solution 1
CI–
▲
+
–
Figure 41 Ion mobility and diffusion of ions through a junction.
82
pH Theory Guide
METTLER TOLEDO
Using the example of sodium and chloride ions we see from the table
and figure above that the sodium and chloride ions diffuse through a
junction from solution 1 into solution 2 at different speeds. Since
Cl– ions in the solution migrate much faster than Na+ ions, a charge
separation occurs.
This charge separation then causes a diffusion potential which counteracts the initial migration. This in turn leads to a dynamic equilibrium
which takes a long time to stabilize. This means that the different diffusion speeds of the ions in the reference electrolyte through the junction
cause a slower response time of the electrode. So it is very important
that the junction is highly porous allowing a strong electrolyte flow in
order that the response time is kept as short as possible.
The charge separation and therefore the diffusion potential Ediff increases when the mobility of the cations and anions is very different.
This effect is particularly noticeable in strongly acidic and basic solutions, the typical solutions often used in pH measurements.
Another factor which determines Ediff is if one of the two solutions is
very dilute. A typical example of such a pH measurement is an ion-deficient sample such as pure water. In this case, the diffusion potential
also increases since the charge difference is amplified by the ion-deficient s ample outside the junction.
To keep the diffusion potential as small as possible you should ensure
that the reference electrolyte is a concentrated and equitransferent solution (equal mobility of anions and cations). This is the case with the
most commonly used KCI and KNO3 reference electrolytes, as can be
seen in Figure 41 on page 82.
pH Theory Guide
METTLER TOLEDO
83
pH Theory Guide
However, despite taking such precautions, the diffusion potential at extreme pH values is considerable even with ideal reference electrolytes.
This is demonstrated in the example below (at 25 °C):
Inner
electrolyte
Sample
solution
Diffusion
potential
∆ pH
KCl (sat.)
HCl (1 mol / L)
Ediff = + 14.1 mV
0.238 pH units
KCl (sat.)
NaOH (1 mol / L) Ediff = – 8.6 mV
0.145 pH units
This description of the diffusion potential makes it clear that some pH
measurements will therefore be more difficult than others. Care should
be taken with very dilute solutions, or solutions which are ion-poor,
such as non-aqueous solutions. In such cases the diffusion potential
will become quite high resulting in an unstable reference signal.
Contaminated junctions also have this effect as the blockage of the
junction inhibits the free flow of electrolyte.
5.5
Calibration / adjustment of the pH measurement setup
There are two settings in the meter which are adapted to the specific
electrode attached to the meter and are affected when the pH electrode
and the meter setup is adjusted, namely the zero point offset (mV) and
the slope (mV / pH) of the electrode. Since there are two settings that
have to be adjusted it follows that a two-point calibration is the minimal adjustment that should be performed.
An adjustment of the zero point and the slope has to be performed to
compensate for any deviations from the theoretical values. These deviations occur due to non-ideal behavior of the electrode. A buffer solution with a pH value of 7.00 corresponds to the zero point of most
glass pH electrodes and is especially intended for the zero point calibration. In most cases, depending on the expected measurement
range, buffer solutions of pH 4.01 or pH 9.21 (or 10.00) are recommended to adjust the slope.
84
pH Theory Guide
METTLER TOLEDO
In the figure below, both these adjustments are illustrated.
The drawing on the left depicts the offset adjustment so that the mV
deviation from the theoretical 0 mV at pH 7.00 is shown. The slope
adjustment is illustrated on the right. Here the deviation from the theoretical 59.16 mV / pH at 25 °C is depicted.
▲
▲
Slope = 59.16 mV/pH
9 mV
pH
pH
Slope = 57.8 mV/pH
▲
▲
mV
▲
▲
7
7
mV
Figure 42 Left: offset adjustment of a pH electrode in the pH meter, right: slope
adjustment of a pH electrode. Solid lines show ideal behavior, dashed lines
show real behavior.
5.6
The influence of temperature on pH measurements
Temperature has an influence on both the electrode and the sample.
We will take a closer look at this influence in the sections below.
5.6.1
Temperature dependence of the electrode
Temperature influences a pH electrode in several different ways:
Slope
Looking at the Nernst equation, which gives the relationship between
measured mV values and pH value of the sample for a pH electrode,
we see that the slope contains the temperature in Kelvin:
RT
E = E0 + 2.3
log aH+
nF
pH Theory Guide
METTLER TOLEDO
85
pH Theory Guide
When we fill in all the numbers, except the temperature in Kelvin (T),
we get:
E = E0 – 0.198 T pH
From this equation we can now clearly see that the slope of an
electrode is linearly dependent on the temperature. Because this dependence is linear the behavior is fully predictable and can be compensated for by a pH meter and electrode with integrated temperature
sensor.
5.6.2
Isothermal intersection
The isothermal intersection depends on the behavior of the individual
potentials E1 to E6 and is a characteristic of every electrode. For an
ideal electrode the calibration lines of different temperatures would intersect at the zero point of the electrode (pH 7.00 / 0 mV) and the slope
would always be proportional to the absolute temperature.
Since the overall potential of the pH electrode is composed of the sum
over E1-E6, which all have their respective temperature dependencies,
the isothermal intersection may not always coincide with the zero point
of the electrode.
It is important for an electrode to have the isothermal intersection and
the zero point as close together as possible, since the nearer these are
to pH 7 the smaller the error in the temperature compensation will be.
The measuring error increases with an increasing temperature difference between the calibration and sample solutions, these errors can be
in the order of 0.1 pH units. The most accurate pH value is obtained
when the temperature of the calibration and sample solutions is identical. These measurement errors are illustrated in Figure 43 below.
86
pH Theory Guide
METTLER TOLEDO
mV
▲
Real isothermal
intersection point
Theoretical isothermal
intersection point
▲
7
14
▲
▲
▲
▲
Measurement error
▲
pH
T1
}
▼
T2
Figure 43 Isothermal intersection, theory and practice.
If the real isothermal intersection does not coincide with the theoretical
one the measurement error can be quite large, depending on the
temperature difference between samples or between sample and calibration. Furthermore, the error can become significant if the real isothermal intersection is very far from the theoretical intersection, and
measurement and calibration differ in temperature.
5.6.3
Further temperature phenomena
The response time of the electrode can also be affected if the temperature changes between or during measurements.
If the change in the temperature of the medium is rapid, a conventional
pH electrode will drift until the temperature of the electrode and the medium becomes equal. In order for a combination electrode to react rapidly to the temperature changes in the sample, the temperature of the
inner pH electrode and the outer reference electrode must always be
identical. This is only possible with a symmetrical arrangement of the
pH and reference elements.
pH Theory Guide
METTLER TOLEDO
87
pH Theory Guide
5.6.4
Temperature dependence of the measured sample
Every sample solution has a characteristic temperature and pH behavior which can be expressed with the so-called temperature coefficient.
This describes how the pH value changes when the temperature
changes. Since this pH change is different for every sample, it is almost impossible to compensate for it.
The first point to note is that the dissociation constant of water itself is
temperature dependent. In pure water when the temperature increases
from 0 and 100 °C, the neutral point shifts 1.34 pH units downwards
as a result of the temperature dependent ion product. In other words
the Kw of water decreases with increasing temperature. A similar behavior is seen in weak acids and bases, since their dissociation constants are also temperature dependent.
The temperature coefficient is determined by two parameters:
• activity coefficient (γ)
• acid constant
The temperature dependence of the activity constant γ becomes larger
when γ is further away from 1, when there is a large deviation between
the concentration and the activity of a solution. This is especially the
case for concentrated solutions and in the presence of ions with a high
electrical charge.
The acid constant pKs is also temperature dependent, but this relationship is non-linear, which means that the dissociation behavior of an
acid changes with temperature. This dissociation behavior causes a
change in the H+ concentration with a change in temperature and thus
a real pH value change.
In general, organic acid / base systems show a higher temperature coefficient than inorganic systems, and alkaline solutions are more temperature dependent than acidic solutions.
88
pH Theory Guide
METTLER TOLEDO
This is illustrated by the following examples:
pH value at:
20 °C
30 °C
0.001 mol / L HCl
3.00
3.00
0.001 mol / L NaOH
11.17
10.83
Phosphate buffer
7.43
7.40
Tris buffer
7.84
7.56
These examples clearly show that large temperature coefficients can
even occur in nearly neutral solutions and therefore that temperature
has to be taken into account when comparing pH measurements obtained at different temperatures. Ideally, samples should be measured
at the same temperature to be able to make comparisons between
them.
In general it is not possible to do temperature compensation for real
changes in pH for chemical solutions. However, temperature compensation tables have been determined for standard buffer solutions.
5.7
Phenomena in the case of special measuring solutions
Different problems may occur when measuring in samples that do not
consist of easy to measure clear, aqueous solutions. These problems
can be of electrical or chemical origin and are briefly discussed in this
section.
5.7.1
Alkaline error
The alkaline effect is the phenomenon where H+ ions in the gel layer of
the pH-sensitive membrane are partly or completely replaced by alkali
ions. This leads to a pH measurement which is too low in comparison
with the number of H+ ions in the sample. Under extreme conditions
where the H+ ion activity can be neglected the glass membrane only
responds to sodium ions.
pH Theory Guide
METTLER TOLEDO
89
pH Theory Guide
Even though the effect is called the alkaline error, it is actually only sodium or lithium ions which cause considerable disturbances. The effect
increases with increasing temperature and pH value (pH > 9), and can
be minimized by using a special pH membrane glass. An example of
electrode behavior under these conditions is given in Figure 44 below.
5.7.2
Acid error
In strongly acidic media, acid molecules are absorbed by the gel layer
leading to a decrease in the H+ ion activity in the gel layer. Conse
quently, an artificially high pH value is registered. The acid error is less
disturbing than the alkaline error and is only relevant at very low pH
values. An illustration of this is also given in Figure 44 below.
mV
▲
Acid error
Theoretical behavior
Experimental
0
Alkaline error
pH
▲
14
Figure 44 Illustration of alkaline and acid error electrode behavior.
5.7.3
Reactions with the reference electrolyte
Another problem source can be the occurrence of chemical reactions
between electrolytes and the measured solution. The resulting precipitates block the pores of the junction and thus increase the electrical resistance considerably.
When using KCI as a reference electrolyte the following ions can precipitate and form compounds of low solubility:
Hg2+, Ag+, Pb2+, CIO4–
90
pH Theory Guide
METTLER TOLEDO
Silver chloride may further react with bromide, iodide, cyanide, and especially with sulfides and sulfide compounds such as cystine and cysteine. Contamination due to silver sulfide results in a black coloration of
the junction. Contamination of the junction may result in unsatisfactory
measurements because of:
• an increase in the response time of the electrode, or
• a diffusion potential (Ediff), which enters into the pH measurement as
a direct error
In order to prevent such reactions between the electrolyte and the sample solution, you can either use an electrolyte which does not react
with the above ions, or you can use an electrode with a double junction and a bridge electrolyte which does not react with the sample.
5.7.4
Organic media
The measurement of pH in organic media or non-aqueous solutions
(less than 5 % water) presents a special challenge, since the classical
definition of pH does not apply for such samples.
When determining the pH value in non-aqueous samples it is important to note that the conventional pH range of pH 0 to pH 14 is based
on the dissociation behavior of water and is therefore not valid. In this
case, the dissociation equilibrium, the ion product of the solvent used
and not the ion product of water is relevant. This can result in completely different concentration ranges for H+ ions in the solvent and thus
a completely d ifferent pH scale. Figure 45 on page 92 illustrates this by
showing the actual valid pH ranges for some common solvents.
pH Theory Guide
METTLER TOLEDO
91
Water
Methanol
Ethanol
Ammonia
Aniline
Diphenylamine
Phenol
0
7
14
Acidic range
21
▲
pH Theory Guide
Acetic acid
28 pH
Alkaline range
Figure 45 pH scale for different solvents.
In applications involving non-aqueous solvents it is common to measure relative rather than absolute pH, e.g. titrations in oil. In this case it
is the potential jump observed when the reaction goes to completion
and not the pH scale that is important. When doing a pH measurement
in a non-aqueous sample it is important to remember that the measurement will not give an absolute pH value. Furthermore, the electrode
will loose its hydrated gel layer around the pH-sensitive membrane. To
ensure that measurements can still be performed you must rehydrate
the gel layer in an ion-rich aqueous solution between experiments.
If you want to measure quantitatively in non-aqueous solvents you can
prepare a calibration curve for the pH glass electrode with different
samples that have a known composition corresponding to the conditions of the samples to be measured. This makes it possible to differentiate the different sample compositions during the measurement,
without having to quantify an absolute value during the measurement.
Remember that non-aqueous solvents are usually very ion-deficient
and that this can result in measurement instabilities.
92
pH Theory Guide
METTLER TOLEDO
5.8
Signal processing
Parameters are measured in a process with the aim of controlling and
influencing the process.
Both the product and the process are important:
Product
– Optimal quality
Process
– Max. yield
– Min. expenditure
– High degree of safety
– Environmental concern
The sensor delivers a signal, which is upgraded and interpreted accordingly. By comparison with the nominal value, measures are taken
which influence the process.
Sensor
Signal
Data acquisition and
interpretation
Valid
signal
Process
Actuation
Action
Figure 46 Typical process control loop.
For pH measurement this means that:.
• The high-resistance sensor signal is transformed into an interferencefree low-resistance signal; preferably with galvanic isolation in order
to prevent ground loops.
• This signal is scaled to pH by calibration, for example by the buffer
values of pH 4 and pH 7.
• Further actions might be taken in order to obtain a valid signal, for
example capturing (recording) the quality parameters of the sensor.
pH Theory Guide
METTLER TOLEDO
93
pH Theory Guide
• Transmission of the scaled pH value in a standardized signal, normally 4 to 20 mA. The current output is insensitive to lead wire resistance and very interference-proof compared to the signal of the
electrode. By the minimal output value of 4 mA (live zero) cable
breaks can be detected.
These tasks are managed by the transmitter. The process can either be
influenced manually (switch function) or automatically via the process
control system.
Base
material
Process
Product
Sensor and
activating
system
Process
control
system
Figure 47 Intersection of the process control system and sensor / activator system.
pH transmitters are often microprocessor controlled devices, which facilitates the operation (e.g. through different calibration methods as
1-point and 2-point calibrations with buffers, 1-point calibration with
process samples, and calibration through setting the parameters of the
electrode which has been tested in the lab). These devices are additionally very flexible in action, since adaptations to the process can be
done through the software.
94
pH Theory Guide
METTLER TOLEDO
pH electrodes have a limited life expectancy and may become useless
for measurements due to contamination. Hence a lot of effort is made
to insure that the pH measurements are reliable and continuous during
the entire process.
For example:
• Redundancy
2 or 3 measuring points at the same measuring site.
➝ Alarm when delta pH > control value
➝ with 3 electrodes: 2 out of 3 conditions: errors not synchronous
• Controlling the electrode with respect to breaks, decay and dynamic
behavior.
• Automatic maintenance and calibration system
In order to exchange failing electrode(s) without interrupting the process, retractable housings have to be used.
Devices for industrial operations are distinguished by their sturdy constructions and high resistance to external influences.
External effects:
• Humidity and dust ➝ IP65 or better
• Ambient temperature ➝ –10 to + 50 °C or better
• Electromagnetic influences are not allowed to affect the functioning
(IEC standards).
The electrode input at the transmitter is characterized by an input offset
resistance, an input current and by the temperature drift behavior.
Under all circumstances the input of the amplifier should be designed
in such a way that the error is not larger than 0.005 pH units. For that
purpose the values of the temperature drift and offset current have to be
limited.
pH Theory Guide
METTLER TOLEDO
95
pH Theory Guide
96
pH Theory Guide
METTLER TOLEDO
Modern amplifiers have the following data:
dE/dT < 25 10 –6V / °C
Rin
> 1012 Ohm
lb
< 10 –12 A
The following example illustrates their influences:
Input resistance RIN
= 1012 Ohm
Input current lb
= 0.5 10 –12 A (25 °C);
lb 2.8 10 –12 A (50 °C)
Temperature drift dE/dT = 20 10 –6 V / °C
Membrane resistance RM = 500 106 Ohm
Influence of the input resistance:
EIN
RM
RIN
EG
RIN
EIN =
∙ EG
RM + RIN
R
error ~ M (pHi – pHa)
RIN
(error in pH units)
for RIN > RM
RM: Resistance of the glass electrode
EG: Potential of the glass electrode
For pHa = 4 (measuring solution):
error 0.002 pH units
Measuring errors which depend on a limited (reduced) input resistance
may be neglected as long as Rin is greater than 1012 Ohm.
pH Theory Guide
METTLER TOLEDO
97
pH Theory Guide
Influence of the input current, if the temperature changes from
25 °C to 50 °C after calibration:
EIN
RM
Ib
EIN = RM Ib
error = 1.15 mV
= 0.018 pH
Influence of the temperature:
dE
E =
T
T = 25 °C
dT
E = 500 10 –6 V = 0.008 pH
Adding all errors: total error = 0.033 pH units
At constant ambient conditions (temperature) the error can be neglected as the change of the temperature is the main disturbing source.
98
pH Theory Guide
METTLER TOLEDO
6
Mathematical parameters
a
b
15
c
7
F
I
pHa
pK
R
S
T
Ediff
Eel
E0
EN
v
z
= Activity of an ion
= Molality (mol / kg solvent)
= Buffer capacity of a solution
= Ion concentration
= Activity coefficient
= Faraday constant (96485 A s mol–1)
= Ion product
= pH value of measured solution
= pH value of inner buffer
= Universal gas constant (8.3143 Nm K–1 mol–1)
= Slope (mV per pH unit)
= Temperature (absolute) (K)
= Diffusion potential
= Electrode potential
= Zero point potential
= Nernst potential
= Velocity of the ion
= Charge of the ion
pH Theory Guide
METTLER TOLEDO
99
www.mt.com/pro
Mettler-Toledo AG
Process Analytics
Phone +41 44 729 62 11
Fax +41 44 729 66 36
Subject to technical changes
© 01/2013 Mettler-Toledo AG
Printed in Switzerland. 30 078 149
For more information