Emerson Model FCL Free Chlorine Measuring System Instruction Manual
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Instruction Manual
PN 51-FCL/rev.H
March 2012
Model FCL
Free Chlorine Measuring System
ESSENTIAL INSTRUCTIONS
WARNINGS
READ THIS PAGE BEFORE PROCEEDING!
RISK OF ELECTRICAL SHOCK
Your purchase from Rosemount Analytical, Inc. has
resulted in one of the finest instruments available for
your particular application. These instruments have
been designed, and tested to meet many national and
international standards. Experience indicates that its
performance is directly related to the quality of the
installation and knowledge of the user in operating
and maintaining the instrument. To ensure their continued operation to the design specifications, personnel
should read this manual thoroughly before proceeding
with installation, commissioning, operation, and maintenance of this instrument. If this equipment is used in
a manner not specified by the manufacturer, the protection provided by it against hazards may be
impaired.
• Failure to follow the proper instructions may cause
any one of the following situations to occur: Loss of
life; personal injury; property damage; damage to
this instrument; and warranty invalidation.
• Ensure that you have received the correct model
and options from your purchase order. Verify that
this manual covers your model and options. If not,
call 1-800-854-8257 or 949-757-8500 to request
correct manual.
• For clarification of instructions, contact your
Rosemount representative.
• Follow all warnings, cautions, and instructions
marked on and supplied with the product.
• Use only qualified personnel to install, operate,
update, program and maintain the product.
• Educate your personnel in the proper installation,
operation, and maintenance of the product.
• Install equipment as specified in the Installation
section of this manual. Follow appropriate local and
national codes. Only connect the product to electrical and pressure sources specified in this manual.
• Use only factory documented components for
repair. Tampering or unauthorized substitution of
parts and procedures can affect the performance
and cause unsafe operation of your process.
• All equipment doors must be closed and protective
covers must be in place unless qualified personnel
are performing maintenance.
• If this equipment is used in a manner not specified
by the manufacturer, the protection provided by it
against hazards may be impaired.
Emerson Process Management
Rosemount Analytical Inc.
2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 757-8500
Fax: (949) 474-7250
http://www.rosemountanalytical.com
© Rosemount Analytical Inc. 2012
Equipment protected throughout by double insulation.
• Installation of cable connections and servicing of this product
require access to shock hazard voltage levels.
• Main power and relay contacts wired to separate power
source must be disconnected before servicing.
• Do not operate or energize instrument with case open!
• Signal wiring connected in this box must be rated at least 240 V.
• Non-metallic cable strain reliefs do not provide grounding
between conduit connections! Use grounding type bushings
and jumper wires.
• Unused cable conduit entries must be securely sealed by
non-flammable closures to provide enclosure integrity in compliance with personal safety and environmental protection
requirements. Unused conduit openings must be sealed with
NEMA 4X or IP65 conduit plugs to maintain the ingress protection rating (NEMA 4X).
• Electrical installation must be in accordance with the National
Electrical Code (ANSI/NFPA-70) and/or any other applicable
national or local codes.
• Operate only with front and rear panels fastened and in place
over terminal area.
• Safety and performance require that this instrument be connected and properly grounded through a three-wire power
source.
• Proper relay use and configuration is the responsibility of the
user.
CAUTION
This product generates, uses, and can radiate radio frequency
energy and thus can cause radio communication interference.
Improper installation, or operation, may increase such interference. As temporarily permitted by regulation, this unit has not
been tested for compliance within the limits of Class A computing devices, pursuant to Subpart J of Part 15, of FCC Rules,
which are designed to provide reasonable protection against
such interference. Operation of this equipment in a residential
area may cause interference, in which case the user at his own
expense, will be required to take whatever measures may be
required to correct the interference.
WARNING
This product is not intended for use in the light industrial,
residential or commercial environments per the instrument’s certification to EN50081-2.
QUICK START GUIDE
FOR FCL ANALYZER
1. Refer to Section 2.0 for installation instructions, and Section 3.0 for wiring instructions.
2. Once connections are secured and verified, apply power to the analyzer.
3. When the analyzer is powered up for the first time, Quick Start screens appear. Using Quick Start is easy.
a. A blinking field shows the position of the cursor.
b. Use the t or u key to move the cursor left or right. Use the p or q key to move the cursor up or down or to
increase or decrease the value of a digit. Use the p or q key to move the decimal point.
c.
Press ENTER to store a setting. Press EXIT to leave without storing changes. Pressing EXIT also returns the
display to the previous screen.
English
Español
Français
>>
5. This screen appears if you have Model FCL-02 (free chlorine and pH). Choose two.
# of sensors?
One
Two
S1 Chlorine Type
free
total
>>
S2 Measure?
pH
Redox
ORP
Temperature in?
°C
4. Choose the desired language. Choose >> to show more choices.
If you have Model FCL-01 (free chlorine only), the screen does not appear.
6. Choose free for sensor 1.
7. Choose pH for sensor 2. If you have Model FCL-01 (free chlorine only), this
screen does not appear.
8. Choose temperature units.
°F
9. The main display appears. The outputs and alarms are assigned to default values.
10. Configure the analyzer for automatic or manual pH correction. Go to the main
menu and choose Program followed by Measurement. Choose Sensor 1 then
Free Chlorine. If you have Model FCL-01 (free chlorine only), choose Manual.
In the next screen, enter the pH of the process liquid. If you have Model FCL-02
(free chlorine and pH), choose Auto.
11. To change outputs, alarms, and temperature-related settings, go to the main
menu and choose Program. Follow the prompts. For a guide to the Program
menu, see the menu tree on the following page.
12. To return the analyzer to the default settings, choose Reset Analyzer in the
Program menu.
MENU TREE FOR CHLORINE/pH MEASUREMENTS
QUICK REFERENCE GUIDE
MODEL FCL
TABLE OF CONTENTS
MODEL FCL
SYSTEM FOR THE DETERMINATION OF FREE CHLORINE
TABLE OF CONTENTS
Section Title
Page
1.0
DESCRIPTION AND SPECIFICATIONS ................................................................
1.1
Features...................................................................................................................
1
1.2
Specifications ...........................................................................................................
2
1.3
Ordering Information ................................................................................................
3
2.0
INSTALLATION .......................................................................................................
4
2.1
Unpacking and Inspection........................................................................................
4
2.2
Installation................................................................................................................
5
3.0
WIRING....................................................................................................................
7
3.1
Power, Alarm, and Output Wiring.............................................................................
7
3.2
Sensor Wiring .........................................................................................................
8
4.0
DISPLAY AND OPERATION ...................................................................................
9
4.1
Display .....................................................................................................................
9
4.2
Keypad.....................................................................................................................
9
4.3
Programming and Calibrating the Analyzer - Tutorial ..............................................
10
4.4
Security ....................................................................................................................
11
4.5
Using Hold ...............................................................................................................
11
5.0
PROGRAMMING THE ANALYZER ........................................................................
12
5.1
General ....................................................................................................................
12
5.2
Changing StartUp Settings.......................................................................................
12
5.3
Configuring and Ranging the Outputs......................................................................
15
5.4
Configuring Alarms and Assigning Setpoints ...........................................................
17
5.5
Selecting Type of Chlorine Measurement ................................................................
20
5.6
Choosing Temperature Units and Manual/Automatic Temperature Compensation .
23
5.7
Setting a Security Code ...........................................................................................
24
5.8
Noise Rejection........................................................................................................
25
5.9
Single Sensor or Dual Sensor Input.........................................................................
25
5.10
Resetting Factory Calibration and Factory Default Settings ....................................
26
5.11
Selecting a Default Screen, Language, and Screen Contrast .................................
26
i
1
MODEL FCL
TABLE OF CONTENTS
TABLE OF CONTENTS CONT’D
Section Title
Page
6.0
CALIBRATION ........................................................................................................
28
6.1
Introduction ..............................................................................................................
28
6.2
Calibrating Temperature...........................................................................................
29
6.3
Calibration - Free Chlorine.......................................................................................
31
6.4
Auto Calibration - pH................................................................................................
36
6.5
Manual Calibration - pH ...........................................................................................
38
6.6
Standardization - pH ................................................................................................
40
6.7
Entering a Known Slope - pH...................................................................................
41
7.0
MAINTENANCE .....................................................................................................
42
7.1
Analyzer ...................................................................................................................
42
7.2
Chlorine Sensor .......................................................................................................
44
7.3
pH Sensor ................................................................................................................
45
7.4
Constant Head Sampler...........................................................................................
46
8.0
TROUBLESHOOTING ............................................................................................
49
8.1
Overview ..................................................................................................................
49
8.2
Troubleshooting Using Fault Codes.........................................................................
49
8.3
Troubleshooting When No Error Message is Showing - Free Chlorine ...................
51
8.4
Troubleshooting When No Error Message is Showing - pH.....................................
54
8.5
Troubleshooting When No Error Message is Showing - General ............................
57
8.6
Simulating Inputs - Chlorine.....................................................................................
57
8.7
Simulating Inputs - pH..............................................................................................
58
8.8
Simulating Temperature ...........................................................................................
59
8.9
Measuring Reference Voltage..................................................................................
60
9.0
RETURN OF MATERIAL ........................................................................................
61
LIST OF TABLES
Number Title
Page
5-1
Default Settings ........................................................................................................
13
7-1
Replacement Parts for FCL (1055-01-11-24-68 or 1055-01-11-24-32-68) ...............
43
7-2
Spare Parts...............................................................................................................
45
7-3
Replacement Parts for Constant Head Flow Controller Assembly (Model FCL-01) .
46
7-4
Replacement Parts for Constant Head Flow Controller Assembly (Model FCL-02) .
48
ii
MODEL FCL
TABLE OF CONTENTS
LIST OF FIGURES
Number Title
Page
2-1
FCL-01 .....................................................................................................................
6
2-2
FCL-02 .....................................................................................................................
6
3-1
Wiring Connections ..................................................................................................
7
3-2
Wiring Diagram for Chlorine Sensor.........................................................................
8
3-3
Wiring Diagram for Chlorine pH Sensor Combination..............................................
8
4-1
Displays During Normal Operation...........................................................................
9
4-2
FCL Keypad..............................................................................................................
9
5-1
Assigning Outputs 1 and 2 ......................................................................................
15
5-2
High Alarm Logic .....................................................................................................
17
5-3
Low Alarm Logic .......................................................................................................
17
6-1
Sensor Current as a Function of Free Chlorine Concentration ................................
31
6-2
Dual Slope Calibration..............................................................................................
34
6-3
Calibration Slope and Offset ....................................................................................
36
7-1
Exploded View of Model FCL Analyzer ....................................................................
43
7-2
Chlorine Sensor Parts ..............................................................................................
45
7-3
Replacement Parts for the Flow Controller Assembly used in Model FCL-01..........
47
7-4
Replacement Parts for the Flow Controller Assembly used in Model FCL-02..........
48
8-1
Pin Out Diagram for Model 499A CL-01-VP Sensor ................................................
50
8-2
Pin Out Diagram for Model 399VP-09 Sensor .........................................................
50
8-3
Simulating Chlorine ..................................................................................................
57
8-4
Simulating pH Inputs ................................................................................................
58
8-5
Three-Wire RTD Configuration.................................................................................
59
8-6
Simulating RTD Inputs..............................................................................................
59
8-7
Checking for a Poisoned Reference Electrode ........................................................
60
iii
About This Document
This manual contains instructions for installation and operation of the Model FCL Free
Chlorine Measuring System.
The following list provides notes concerning all revisions of this document.
Rev. Level
Date
A
2/03
This is the initial release of the product manual. The manual
has been reformatted to reflect the Emerson documentation
style and updated to reflect any changes in the product offering.
B
4/03
Updated CE certification
C
6/03
Revisions throughout to reflect product modification.
D
12/03
Add text to Section 8.1
E
10/04
Revised dimensional and exploded view drawings.
F
1/05
Revise low flow cell wetted materials specs
G
1/06
Revised wetted parts specs
H
03/12
Update addresses - mail and web, and DNV certificate logo
Emerson Process Management
Liquid Division
2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 757-8500
Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2006
Notes
MODEL FCL
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
SECTION 1.0.
DESCRIPTION AND SPECIFICATIONS
1.1 FEATURES
1.2 SPECIFICATIONS
1.3 ORDERING INFORMATION AND ACCESSORIES
• COMPLETE SYSTEM INCLUDES sensor, connecting cable, analyzer, and flow controller
• CONTINUOUS pH CORRECTION eliminates expensive and messy reagents and troublesome sample conditioning systems
• MEASURES FREE CHLORINE IN SAMPLES having pH as high as 9.5 1
• VARIOPOL QUICK-DISCONNECT FITTINGS make replacing sensors easy
• FEATURE-PACKED ANALYZER: dual outputs, three fully-programmable alarm relays, twoline display
1
In some cases, the sensor can be used in samples having pH as great as 10.0. Consult the factory.
1.1 FEATURES
The FCL free chlorine system is intended for the determination of free chlorine in fresh water. Unlike free chlorine analyzers from other manufacturers, the FCL does not use
expensive sample conditioning systems or messy reagents
to control pH. Instead, the analyzer automatically compensates for changes in the pH of the sample. The FCL is not
intended for the determination of total chlorine or combined
chlorine (like monochloramine). Nor, can the FCL be used for
the determination of chlorine in seawater.
APPLICATIONS
ing by measuring the pH and applying a correction to the
raw chlorine sensor signal. The correction is valid between
pH 6.0 and 9.5. For samples having pH between 9.5 and
10.0, consult the factory.
The FCL is available in two options: Model FCL-01 with
manual pH correction and Model FCL-02 with continuous pH
correction. Choose the FCL-01 if the pH varies less than 0.2
or if pH changes are predictable or seasonal. Choose the
FCL-02 if the pH varies more than 0.2. To provide the continuous pH correction, the FCL-02 requires a separate pH sensor.
The Model FCL uses a membrane-covered amperometric
sensor. A polarizing voltage applied to a platinum cathode
behind the membrane reduces the chlorine diffusing through
the membrane and keeps the concentration of chlorine in
the sensor equal to zero. The current generated by the cathode reaction is proportional to the rate of diffusion of chlorine through the membrane. Because the concentration of
chlorine in the sensor is zero, the diffusion rate and the current are proportional to the concentration of chlorine in the
sample.
Maintenance is fast and easy. Replacing a membrane
requires no special tools or fixtures. A screw cap holds the
pre-tensioned membrane in place. Replacing the electrolyte
solution takes only minutes.
There is a difficulty, however. Amperometric free chlorine
sensors measure only hypochlorous acid. Because free
chlorine is a pH-dependent mixture of hypochlorous acid
and hypochlorite ion, a change in pH will cause the sensor
response to change even though the free chlorine level
remained constant. Most manufacturers solve the problem
by treating the sample with acid, which lowers the pH and
converts hypochlorite ion into hypochlorous acid. The FCL
avoids the expense and inconvenience of sample condition-
Valves, rotameters, and pressure regulators to control sample flow are things of the past with the Model FCL. A constant head overflow sampler ensures the correct sample
flow to each sensor. To eliminate wiring hassles, quick-disconnect Variopol cable is standard.
The FCL includes the easy-to-use 1055 analyzer. The analyzer features two fully programmable 4-20 mA outputs and
three fully programmable relays. The back-lit, two line display allows the user to read sample pH and chlorine concentration at a glance.
Stable free chlorine standards do not exist. The chlorine
sensor must be calibrated using the results of a laboratory
test run on a grab sample.
1
MODEL FCL
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.2 SPECIFICATIONS — GENERAL
SPECIFICATIONS — ANALYZER
Sample requirements:
Case: ABS, NEMA 4X
Pressure: 3 to 65 psig (122 to 549 kPa abs)
A check valve in the inlet prevents the sensor
flow cells from going dry if sample flow is lost.
The check valve opens at 3 psig (122 kPa abs).
If the check valve is removed, minimum pressure is 1 psig (108 kpa abs).
Temperature: 32 to 122°F (0 to 50°)
Minimum Flow: 3 gal/hr (11 L/hr)
Maximum flow: 80 gal/hr (303 L/hr); high flow
causes the overflow tube to back up
Sample Conductivity: >50 µS/cm at 25°C
Process connection: 1/4-in OD tubing compression
fitting (can be removed and replaced with barbed
fitting for soft tubing)
Display: Two-line, 16-character, back-lit. Character
height: 4.8 mm.
Languages: English, German, Italian, Spanish,
French, Portuguese
Ambient temperature and humidity: 0 to 50°C (32
to 122°F); RH 5 to 95% (con-condensing)
The analyzer can be operated between -20 and
60°C (-4 to 140°F) with some degradation in display performance.
Power: 115/230Vac ± 15%, 50/60 Hz ± 6%, 8.0 W.
Installation category II.
Equipment protected throughout by double insulation.
Ordinary Location:
Drain connection: 3/4-in barbed fitting. Sample must
drain to open atmosphere
Wetted parts:
Overflow sampler and flow cell: acrylic, polycarbonate, Kynar®1, nylon, silicone
Chlorine sensor: Noryl®2, Viton®3, wood, silicone,
polyethersulfone, polyester, and platinum
pH sensor: Tefzel®4, Viton, glass, ceramic
Response time to step change in chlorine concentration: <80 sec to 95% of final reading for inlet
sample flow of 3 gph (11 L/hr)
Weight/shipping weight:
Model FCL-01: 10 lb/13 lb (4.5 kg/6.0 kg)
Model FCL-02: 11 lb/14 lb (5.0 kg/6.5 kg)
[rounded to the nearest 1 lb. (0.5 kg)]
SPECIFICATIONS — SENSOR
Free chlorine range: 0 to 10 ppm as Cl2. For higher
ranges, consult the factory.
pH correction range: 6.0 to 9.5. For samples having
pH between 9.5 and 10.0, consult the factory.
For manual pH correction, choose option -01.
For continuous pH correction choose option -02.
12RN
POLLUTION DEGREE 2: Normally only non-conductive pollution occurs. Occasionally, however, a
temporary conductivity caused by condensation
must be expected.
RFI/EMI:
EN-61326
LVD:
EN-61010-1
Outputs: Two 4-20 mA or 0-20 mA isolated outputs.
Continuously adjustable. Linear or logarithmic.
Maximum load 600 ohms. Output dampening with
time constant of 5 sec is user-selectable.
Alarms: Three alarm relays for process measurement(s) or temperature. Alarm 3 can be
configured as a fault alarm, instead of a
process alarm. Each relay can be configured
independently. Alarm logic (high or low activation)
and deadband are user-programmable.
Relays: Form C, single pole double throw, epoxy sealed
28 Vdc
Resistive
5.0 A
Inductive
3.0 A
115 Vac
5.0 A
3.0 A
Accuracy: Accuracy depends on the accuracy of the
chemical test used to calibrate the sensor.
1 Kynar is a registered trademark of Elf Atochem North America.
2 Noryl is a registered trademark of General Electric.
Electrolyte volume: 25 mL (approx.)
3 Viton is a registered trademark of E.I. duPont de Nemours & Co.
4 Tefzel is a registered trademark of E.I. duPont de Nemours & Co.
Electrolyte life: 3 months (approx.); for best results
replace electrolyte monthly.
2
MODEL FCL
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.3 ORDERING INFORMATION
FCL Free Chlorine Measuring System. The FCL is a complete system for the determination of free chlorine in
aqueous samples. It consists of the sensor(s), analyzer, and constant head overflow cup to control sample flow.
All components are mounted on a backplate. Model option -02 includes a pH sensor for continuous, automatic pH
correction. Three replacement membranes and a 4-oz. bottle of electrolyte solution are shipped with each sensor.
FCL
FREE CHLORINE MEASURING SYSTEM
CODE
01
02
pH CORRECTION (required selection)
Without continuous pH correction
With continuous pH correction
FCL-02
EXAMPLE
COMPONENT PARTS
ANALYZER MODEL
1055-01-11-24-68
1055-01-11-24-32-68
DESCRIPTION
1055 analyzer, single input (chlorine), wall mount, 115/230 Vac
1055 analyzer, dual input (chlorine and pH), wall mount, 115/230 Vac
SENSOR MODEL
499ACL-01-54-VP
399VP-09-305
DESCRIPTION
Free chlorine sensor with Variopol connector
pH sensor with Variopol connector
SENSOR CABLE
23747-04
23645-08
DESCRIPTION
Interconnecting cable, Variopol for 499ACL sensor, 4 ft
Interconnecting cable, Variopol for 399VP sensor, 4 ft
ACCESSORIES
PART #
9240048-00
DESCRIPTION
Tag, stainless steel (specify marking)
3
MODEL FCL
SECTION 2.0
INSTALLATION
SECTION 2.0.
INSTALLATION
2.1 UNPACKING AND INSPECTION
2.2 INSTALLATION
2.1 UNPACKING AND INSPECTION
Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions. Save the box. If
there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present. If
items are missing, notify Rosemount Analytical immediately.
2.1.1 FCL-01 (free chlorine without continuous pH correction)
The FCL-01 consists of the following items mounted on a back plate.
1. Model 1055-01-13-24 analyzer with sensor cable attached.
2. Constant head overflow sampler with flow cell for chlorine sensor.
The free chlorine sensor (Model 499ACL-01-54-VP), three membrane assemblies, and a bottle of electrolyte solution are in a separate package.
2.1.2 FCL-02 (free chlorine with continuous pH correction)
The FCL-02 consists of the following items mounted on a back plate.
1. Model 1055-01-13-24-32 analyzer with sensor cables attached.
2. Constant head overflow sampler with flow cells for pH and chlorine sensors.
3. Stand to hold pH buffer solution during calibration.
The free chlorine sensor (499ACL-01-54-VP), shipped with three membrane assemblies and a bottle of electrolyte
solution, and the 399VP-09-305 pH sensor are in separate packages.
4
MODEL FCL
SECTION 2.0
INSTALLATION
2.2 INSTALLATION
2.2.1 General Information
1. Although the system is suitable for outdoor use, do not install it in direct sunlight or in areas of extreme temperature.
2. To keep the analyzer enclosure watertight, install plugs (provided) in the unused cable openings.
3. Install the system in an area where vibrations and electromagnetic and radio frequency interference are minimized or absent.
4. Be sure there is easy access to the analyzer and sensors.
2.2.2 Sample Requirements
Be sure the sample meets the following requirements:
1. Temperature: 32 to 122ºF (0 to 50ºC)
2. Pressure: 3 to 65 psig (122 to 549 kPa abs)
3. Minimum flow: 3 gal/hr (11 L/hr)
2.2.3 Mounting, Inlet, and Drain Connections
The Model FCL is intended for wall mounting only. Refer to Figure 2-1 or 2-2 for details. The sensor(s) screw into
the flow cell adapters as shown in the figures. For Model FCL-02 (free chlorine with continuous pH adjustment),
the pH sensor must be installed as shown in Figure 2-2.
A 1/4-inch OD tubing compression fitting is provided for the sample inlet. If desired, the compression fitting can
be removed and replaced with a barbed fitting. The fitting screws into a 1/4-inch FNPT check valve. The check
valve prevents the sensor flow cells from going dry if sample flow is lost.
The sample drains through a 3/4-inch barbed fitting. Attach a piece of soft tubing to the fitting and allow the waste
to drain open atmosphere. Do not restrict the drain line.
Remove the foam packing insert between the outer tube and the inner overflow tube. Adjust the sample flow
until the water level is even with the central overflow tube and excess water is flowing down the tube.
2.2.4 Electrical Connections
Refer to Section 3.1 for details.
2.2.5 Installing the Sensor(s)
The Model FCL is provided with sensor cables pre-wired to the analyzer. Connect the chlorine sensor (Model
499ACL-01-54-VP) to the cable labeled CL. Connect the pH sensor (Model 399VP-09-305) to the cable labeled
pH. The terminal end of the sensor is keyed to ensure proper mating with the cable receptacle. Once the key has
slid into the mating slot, tighten the connection by turning the knurled ring clockwise.
The sensor(s) screw into the plastic fitting(s), which are held in the flow cell(s) by the union nut. Do not remove the
protective cap on the sensor(s) until ready to put the sensor(s) in service.
5
MODEL FCL
SECTION 2.0
INSTALLATION
INCH
MILLIMETER
FIGURE 2-1. Model FCL-01
INCH
MILLIMETER
FIGURE 2-2. Model FCL-02
6
MODEL FCL
SECTION 3.0
WIRING
SECTION 3.0.
WIRING
3.1 POWER, ALARM, AND OUTPUT WIRING
3.2 SENSOR WIRING
3.1 POWER, ALARM, AND OUTPUT WIRING
See Figure 3-1 for identification of power, alarm, and output terminals. Note that the sensors are already wired to
the analyzer.
AC power wiring should be 14 gauge or greater. Run the power wiring through the conduit opening nearest the
power terminal (TB1). Provide a switch or breaker to disconnect the analyzer from the main power supply. Install
the switch or breaker near the analyzer and label it as the disconnecting device for the analyzer.
Keep output signal wiring separate from power wiring. Do run signal and power wiring in the same conduit or close
together in a cable tray.
For best EMI/RFI protection use shielded output signal cable enclosed in an earth-grounded metal conduit.
Connect the shield to earth ground at TB1-4.
Keep output wiring at least one foot from high voltage conductors.
To reduce stress on the wiring connections, do not remove the hinged front panel from the base while installing
wiring. Be sure the leads are sufficiently long to avoid stress on the conductors.
WARNING:
RISK OF ELECTRICAL SHOCK
AC connections and grounding must be
in compliance with UL 508 or local electrical code. DO NOT apply power to the
analyzer until all electrical connections
are verified and secure.
Figure 3-1. Wiring connections.
7
MODEL FCL
SECTION 3.0
WIRING
3.2 SENSOR WIRING
The Model FCL is provided with sensor cables pre-wired to the analyzer.
If it is necessary to replace the cable, refer to the wiring diagrams below. Figure 3-2 is the sensor wiring diagram
for Model FCL-01 (free chlorine sensor only). Figure 3-3 is the sensor wiring diagram for Model FCL-02 (free chlorine and pH sensor). The jumper (PN 23980-00) between TB3-1 (anode) and TB5-6 (pH reference) in Figure 3-3
is an integral part of the circuit. It must be installed as shown.
Figure 3-2. Wiring diagram for chlorine sensor.
Figure 3-3. Wiring diagram for
chlorine pH sensor combination.
8
MODEL FCL
SECTION 4.0
DISPLAY AND OPERATION
SECTION 4.0
DISPLAY AND OPERATION
4.1
4.2
4.3
4.4
4.5
DISPLAY
KEYPAD
PROGRAMMING AND CALIBRATING THE ANALYZER - TUTORIAL
SECURITY
USING HOLD
4.1. DISPLAY
The Model FCL analyzer has a two-line
display. The display can be customized
to meet user requirements (see Section
5.11). Figure 4-1 shows some of the displays. View A is the default screen for
Model FCL-02 (chlorine and pH sensor).
View C is the default screen for Model
FCL-01 (chlorine sensor only).
The FCL analyzer has information
screens that supplement the data in the
main display. Press " or ' to view the
information screens. The last information screen is the software version.
During calibration and programming,
key presses cause different displays to
appear. The displays are self-explanatory and guide the user step-by-step
through the procedure.
FIGURE 4-1. Displays During Normal Operation
Screen A shows chlorine and pH. The temperature shown is the temperature measured by the chlorine sensor. Screen B shows chlorine and pH
and the temperature measured by each sensor. Screen C shows the data
for the chlorine sensor only.
4.2 KEYPAD
Figure 4-2 shows the keypad.
FIGURE 4-2. FCL Analyzer Keypad
Four arrow keys move the cursor around the screen. A blinking word or
numeral show the position of the cursor. The arrow keys are also used to
change the value of a numeral. Pressing ENTER stores numbers and settings and moves the display to the next screen. Pressing EXIT returns to
the previous screen without storing changes. Pressing MENU always
causes the main menu screen to appear. Pressing MENU followed by
EXIT causes the main display to appear.
9
MODEL FCL
SECTION 4.0
DISPLAY AND OPERATION
4.3 PROGRAMMING AND CALIBRATING THE ANALYZER TUTORIAL
Setting up and calibrating the FCL is easy. The following tutorial describes
how to move around in the programming menus. For practice, the tutorial also
describes how to assign chlorine values to the 4 and 20 mA outputs for sensor 1 (free chlorine sensor).
Calibrate
Program
Calibrate
Hold
Display
Hold
Program
Display
Outputs
Alarms
Measurement
>>
1. If the MENU screen (shown at the left) is not already showing, press
MENU. Calibrate is blinking, which means the cursor is on Calibrate.
2. To assign values to current outputs, the Program sub-menu must be open.
Press q. The cursor moves to Program (Program blinking). Press ENTER.
Pressing ENTER opens the Program sub-menu.
3. The Program sub-menu permits the user to set outputs, alarms, automatic or manual temperature compensation, and a security code. When
the sub-menu opens, Outputs is blinking, which means the cursor is on
Outputs. Press q or u (or any arrow key) to move the cursor around the
display. Move the cursor to >> and press ENTER to cause a second
screen with more program items to appear. There are three screens in the
Program menu. Pressing >> and ENTER in the third screen causes the
display to return to the first screen (Outputs, Alarms, Measurement).
4. For practice, assign values to the 4 and 20 mA outputs for sensor 1. Move
the cursor to Outputs and press ENTER.
Output Range
5. The screen shown at left appears. The cursor is on Output Range (blinking). Output range is used to assign values to the low and high current
outputs. Press ENTER.
Output Configure
Output Range?
Output1
Output2
Out1 S1 Range?
4mA
00.00ppm
6. The screen shown at left appears. The FCL has two outputs, output 1 and
output 2. Move the cursor to the desired output and press ENTER. For
purposes of the example, choose Output 1.
7. The screen shown at left appears. Out1 S1 in the top line means output
1 (Out1) is assigned to sensor 1 (S1). Either output can be assigned to
either sensor (sensor and output assignments are made under the
Output Configure menu shown in step 5). Use the Out1 S1 Range?
screen to assign a chlorine concentration to the 4 mA output.
a. Use the arrow keys to change the concentration to the desired value.
Press t or u to move the cursor from digit to digit. Press p or q to
increase or decrease the value of the digit. Holding p or q down
causes the numeral to continuously scroll up or down.
b. To move the decimal point, press t or u until the cursor is on the
decimal point. Press p to move the decimal point to the right. Press
q to move the decimal point to the left.
c.
Out1 S1 Range?
20mA
20.00ppm
Output Range?
Output1
Output2
Press ENTER to store the setting.
8. The screen shown at left appears. Use this screen to assign a full scale
chlorine concentration to the 20 mA output. Use the arrow keys to change
the chlorine to the desired value. Press ENTER to store the setting.
9. The screen shown at left appears. To assign values to the low and high
currents for output 2, select Output 2 and follow the prompts.
10. To return to the main menu, press MENU. To return to the main display
press MENU then EXIT, or press EXIT repeatedly until the main display
appears. To return to the previous display press EXIT.
NOTE
To store values or settings, press ENTER before pressing EXIT.
10
MODEL FCL
SECTION 4.0
DISPLAY AND OPERATION
4.4 SECURITY
4.4.1 How the Security Code Works
Use the security code to prevent accidental or unwanted changes to program settings, displays, and calibration.
Enter Security
Code
000
1. If a security code has been programmed, pressing MENU causes the
security screen to appear.
2. Enter the three-digit security code.
Invalid Code
3. If the entry is correct, the main menu screen appears. If the entry is incorrect, the Invalid Code screen appears. The Enter Security Code screen
reappears after 2 seconds.
4.4.2 Bypassing the Security Code
Enter 555. The main menu will open.
4.4.3 Setting a Security Code
See Section 5.6.
4.5 USING HOLD
4.5.1 Purpose
The analyzer output is always proportional to measured pH or chlorine. To prevent unwanted alarms and improper operation of control systems or dosing pumps, place the analyzer in hold before removing the sensor for calibration and maintenance. Be sure to remove the analyzer from hold once caliHold
bration is complete. During hold, both outputs remain at the last value. Once in
hold, the analyzer remains there indefinitely. While in hold, the screen
shown to the left appears periodically.
4.5.2 Using the Hold Function
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
Hold
Calibrate
Program
Display
Hold Outputs and
Alarms?
1. Press MENU. The main menu screen appears. Choose Hold.
Yes
No
2. The Hold Outputs and Alarms ? screen appears. Choose Yes to place
the analyzer in hold. Choose No to take the analyzer out of hold.
3. The main display screen will appear.
11
MODEL FCL
SECTION 5.0
PROGRAMMING THE ANALYZER
SECTION 5.0
PROGRAMMING THE ANALYZER
5.1
5.2
5.3
5.4
5.5
5.6
GENERAL
CHANGING STARTUP SETTINGS
CONFIGURING AND RANGING THE OUTPUTS
CONFIGURING ALARMS AND ASSIGNING SETPOINTS
SELECTING THE TYPE OF CHLORINE MEASUREMENT
CHOOSING TEMPERATURE UNITS AND MANUAL OR AUTOMATIC TEMPERATURE COMPENSATION
5.7 SETTING A SECURITY CODE
5.8 NOISE REJECTION
5.9 SINGLE SENSOR OR DUAL SENSOR INPUT
5.10 RESETTING FACTORY CALIBRATION AND FACTORY DEFAULT SETTINGS
5.11 SELECTING A DEFAULT SCREEN, LANGUAGE, AND SCREEN CONTRAST
5.1 GENERAL
This section describes how to do the following:
1. configure and assign values to the current outputs
2. configure and assign setpoints to the alarm relays
3. choose the type of chlorine measurement being made
4. choose temperature units and manual or automatic temperature mode
5. set a security code
6. tell the analyzer the frequency of the ac power (needed for optimum noise rejection)
7. tell the analyzer the number of sensors being used
8. reset the analyzer to factory calibration and default settings
9. select a default display screen
Default settings are shown in Table 5-1 on the following page. To change a default setting, refer to the section listed in the table. To reset default settings, see Section 5.10.
5.2 CHANGING STARTUP SETTINGS
When the Model FCL analyzer is powered up for the first time, Quick Start screens appear, which enable the user
to quickly configure the analyzer for free chlorine (Model FCL-01) or for free chlorine and pH (Model FCL-02).
Because the analyzer can be used to measure other chlorine compounds, it must be specifically configured to measure free chlorine. If incorrect settings were entered at startup, enter the correct settings now. Refer to Section 5.9 to
change the number of sensors. Refer to Section 5.5 to configure the analyzer to measure free chlorine and pH.
FOR BEST RESULTS, ENTER THE NUMBER OF SENSORS BEING USED (SECTION 5.9), AND IDENTIFY FREE CHLORINE FOR SENSOR 1 AND pH FOR SENSOR 2 (SECTION 5.5) BEFORE MAKING OTHER PROGRAM SETTINGS.
12
MODEL FCL
SECTION 5.0
PROGRAMMING THE ANALYZER
TABLE 5-1. DEFAULT SETTINGS
1. SENSOR-OUTPUT ASSIGNMENTS
Model
Output 1
Output 2
Section
FCL-01
chlorine
temperature
5.3 and 5.9
FCL-02
chlorine (sensor 1)
pH (sensor 2)
5.3 and 5.9
2. OTHER OUTPUT SETTINGS
Output
Dampening
0 or 4 mA
Mode
Section
1
off
4
Linear
5.3
2
off
4
Linear
5.3
3. OUTPUT RANGES
Measurement
Range
Section
free chlorine - ppm
0 to 20 ppm
5.3
pH
0 to 14
5.3
Temperature
0 to 100°C
5.3
4. ALARM CONFIGURATION
1
Alarm
2
3
If AL3 is a
sensor alarm
Section
FCL-01
chlorine
temperature
fault
chlorine
5.4
FCL-02
chlorine
pH
fault
temperature
5.4
High or low
high
high
NA
high
5.4
High
Low
Deadband
Section
20.00
0.00
0.00
5.4
pH
14.00
0.00
0.00
5.4
Temperature
100.0
0.0
0.0
5.4
5. ALARM SETPOINTS
Chlorine
13
MODEL FCL
SECTION 5.0
PROGRAMMING THE ANALYZER
TABLE 5-1. DEFAULT SETTINGS (continued)
6. TEMPERATURE RELATED SETTINGS
Section
Units
°C
5.6
Automatic temperature compensation (chlorine)
On
5.6
Automatic temperature compensation (pH)
On
5.6
7. MISCELLANEOUS SETTINGS
Section
Language
English
5.11
Hold
off
4.5
Security code
000 (no security code)
5.7
ac power frequency
60 Hz
5.8
14
MODEL FCL
SECTION 5.0
PROGRAMMING THE ANALYZER
5.3 CONFIGURING AND RANGING THE OUTPUTS.
5.3.1 Purpose
The Model FCL analyzer has two current outputs. This section describes how to configure and range the outputs.
CONFIGURE THE OUTPUTS FIRST.
1. Configuring an output means
a. Selecting either a 4-20 mA or 0-20 mA output,
b. Assigning a sensor and a measurement (free chlorine or pH) to output 1 and output 2,
c.
Turning on or turning off output current dampening,
d. Choosing a linear or logarithmic output.
2. Ranging the outputs means assigning values to the low (0 or 4 mA) and high (20 mA) outputs.
5.3.2 Definitions
1. CURRENT OUTPUTS. The analyzer provides either a continuous 4-20 mA or 0-20 mA output current directly
proportional to chlorine concentration or pH.
2. ASSIGNING OUTPUTS. Figure 5-1 shows the ways in which the outputs can be assigned.
3. DAMPEN. Output dampening smooths out noisy readings. It also increases the response time of the output.
With output dampening the time to reach 63% of final reading following a step change is 5 sec. Output dampening does not affect the response time of the display.
4. MODE. The current output can be made directly proportional to the displayed value (linear mode) or directly
proportional to the common logarithm of the displayed value (log mode).
FIGURE 5-1. Assigning Outputs 1 and 2
15
MODEL FCL
SECTION 5.0
PROGRAMMING THE ANALYZER
5.3.3. Procedure: Configure Outputs.
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
Calibrate
Hold
Program
Display
Outputs
Alarms
Measurement
1. Press MENU. The main menu screen appears. Choose Program.
2. Choose Outputs.
>>
3. Choose Output Configure.
Output Range
Output Configure
4. Choose Output1 or Output2.
Output Config?
Output1
Output2
OutM is for?
Sensor1
Sensor2
OutM is for?
Measurement
Temp
5. Choose Sensor1 (chlorine) or Sensor2 (pH). Either sensor can be
assigned to either output.
6. Choose Measurement or Temp. If the output selected was assigned to
Sensor 1, Measurement means chlorine. If the output selected was
assigned to Sensor 2, Measurement means pH.
7. Make the appropriate settings:
a. Choose 4-20 mA or 0-20 mA.
b. Choose Yes or No for output dampening.
c.
Choose Linear or Log output.
8. The display returns to the Output Config? screen. Select the other output or press EXIT to return to the previous screen. To return to the main
display, press MENU followed by EXIT.
5.3.4. Procedure: Assigning Values to the Low and High Current Outputs (Output Ranging)
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
Calibrate
Hold
Program
Display
Outputs
Alarms
Measurement
1. Press MENU. The main menu screen appears. Choose Program.
2. Choose Outputs.
>>
Output Range
3. Choose Output Range. Choose Output1 or Output2.
Output Configure
4. Make the appropriate settings.
a. Assign a value to the low current (0 mA or 4 mA) output.
b. Assign a value to the high current (20 mA) output.
5. The display returns to the Output Range screen. Select the other output
or press EXIT to return to the previous screen. To return to the main display, press MENU followed by EXIT.
16
MODEL FCL
SECTION 5.0
PROGRAMMING THE ANALYZER
5.4 CONFIGURING ALARMS AND ASSIGNING SETPOINTS
5.4.1 Purpose
This section describes how to do the following:
1. assign an alarm relay to a sensor,
2. set the alarm logic to high or low,
3. assign values to the alarm setpoints,
4. set the alarm deadbands.
ALARM RELAYS MUST BE CONFIGURED BEFORE ASSIGNING SETPOINTS.
5.4.2 Definitions
1. ASSIGNING ALARMS. There are three alarms (AL1, AL2, and AL3). Alarms 1 and 2 can be assigned to either
sensor. For example, AL1 and AL2 can be assigned to sensor 1 with, perhaps, one alarm configured as a high
alarm and the other as a low alarm, and AL3 can be assigned to sensor 2. Alarm 3 can be assigned to either
sensor or used as a fault alarm. The fault alarm activates when a fault exists in a sensor or the analyzer.
2. FAULT ALARM. A fault condition exists when the Model FCL analyzer detects a problem with a sensor or with
the analyzer that is likely to cause seriously erroneous readings. If Alarm 3 was programmed as a fault alarm,
the alarm 3 relay will activate. The word Fault will appear alternately in the display with the reading.
3. ALARM LOGIC, SETPOINTS, AND DEADBANDS. See Figures 5-2 and 5-3.
FIGURE 5-2. High Alarm Logic
FIGURE 5-3. Low Alarm Logic
The alarm activates when the pH exceeds the high
setpoint. The alarm remains activated until the reading
drops below the value determined by the deadband.
The alarm activates when the chlorine concentration
drops below the low setpoint. The alarm remains activated until the reading increases above the value determined by the deadband.
Alarm relays are single pole-double throw (SPDT). When an alarm is activated, the coil is energized.
When an alarm activates, AL1, AL2, or AL3 (as appropriate) appears periodically in the display.
17
MODEL FCL
SECTION 5.0
PROGRAMMING THE ANALYZER
5.4.3 Procedure: Configuring Alarms
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
Calibrate
Hold
Program
Display
Outputs
Alarms
Measurement
1. Press MENU. The main menu screen appears. Choose Program.
2. Choose Alarms.
>>
3. Choose Alarm Configure.
Alarm Setpoints
Alarm Configure
4. Choose Alarm 1 (AL1), Alarm 2 (AL2), or Alarm 3 (AL3).
Alarm Config?
AL1
AL2
AL3
5. For AL1 or AL2
AL1 is for?
Sensor1
a. Choose Sensor 1 (chlorine) or Sensor 2 (pH).
Sensor2
AL1 S1 is for?
Measurement
Temp
b. Choose Measurement or Temp.
c.
Choose High or Low.
d. Set the alarm Deadband.
6. The display returns to the Alarm Configure? screen. Select another
alarm or press EXIT to return to the previous screen. To return to the main
display, press MENU followed by EXIT.
7. For AL3
AL3 is for?
Sensor1
Fault
Sensor2
a. Choose Sensor1 (chlorine), Sensor2 (pH), or Fault.
b. For sensor 1 or 2, choose Measurement or Temp.
c.
Choose High or Low. Set the deadband.
d. Choosing Fault means AL3 will activate when a sensor or analyzer
fault exists. There is no user setting to make.
8. The display returns to the Alarm Configure? screen. Select another
alarm or press EXIT to return to the previous screen. To return to the main
display, press MENU followed by EXIT.
18
MODEL FCL
SECTION 5.0
PROGRAMMING THE ANALYZER
5.4.4 Procedure: Programming Alarm Setpoints
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
Calibrate
Hold
Program
Display
Outputs
Alarms
Measurement
1. Press MENU. The main menu screen appears. Choose Program.
2. Choose Alarms.
>>
Alarm Setpoints
3. Choose Alarm Setpoints.
Alarm Configure
4. Choose Alarm 1 (AL1), Alarm 2 (AL2), or Alarm 3 (AL3).
Select Alarm?
AL1
AL2
AL3
AL1 S1 Setpoint?
High
20.00ppm
5. The display shows the alarm selected (AL1) and the configuration. The
alarm is for Sensor 1 (S1), and the logic is high. Use the arrow keys to
change the alarm setpoint.
6. The display returns to the Select Alarm? screen. Select another alarm or
press EXIT to return to the previous screen. To return to the main display,
press MENU followed by EXIT.
19
MODEL FCL
SECTION 5.0
PROGRAMMING THE ANALYZER
5.5 SELECTING THE TYPE OF CHLORINE MEASUREMENT
5.5.1 Purpose
This section describes how to do the following:
1. Program the analyzer to measure free chlorine (and pH). This step is necessary because the Model FCL analyzer can be used with other sensors to measure other chlorine oxidants. It can also be used to measure ORP
(oxidation reduction potential). When used in the Model FCL, the analyzer should be programmed to measure
either free chlorine (FCL-01) or free chlorine and pH (FCL-02).
2. Set automatic or manual pH correction for the free chlorine measurement
3. Determine the level of electronic filtering of the sensor current
4. Enable or disable dual slope calibration
5. Make various pH measurement settings. The analyzer supplied with the Model FCL is designed to be as versatile as possible. The pH settings listed below are needed in some applications, but are not used when pH is
measured for correcting free chlorine readings.
a. solution temperature correction
b. analyzer isopotential point
c. enable or disable glass impedance fault.
The settings should be left at their default values.
5.5.2 Definitions — Chlorine
1. FREE CHLORINE. Free chlorine is the result of adding sodium hypochlorite (bleach), calcium hypochlorite
(bleaching powder), or chlorine gas to fresh water. Free chlorine is the sum of hypochlorous acid (HOCl) and
hypochlorite ion (OCl-).
2. TOTAL CHLORINE. Total chlorine is the sum of free and combined chlorine. Combined chlorine generally
refers to chlorine oxidants in which chlorine is combined with ammonia or organic amines.
3. MONOCHLORAMINE. Monochloramine (NH2Cl) is the product of a chemical reaction between ammonia and
chlorine. It is commonly used to disinfect drinking water.
4. pH CORRECTION. Free chlorine is the sum of hypochlorous acid (HOCl) and hypochlorite ion (OCl-). The relative amount of each depends on pH. As pH increases, the concentration of HOCl decreases and the concentration of OCl- increases. Because the sensor responds only to HOCl, a pH correction is necessary to properly convert the sensor current into a free chlorine reading. The FCL uses either automatic or manual pH correction. In automatic pH correction (Model FCL-02), the analyzer continuously monitors the pH of the sample
and corrects the free chlorine reading for changes in pH. In manual pH correction, the user must enter the pH
of the sample. Generally, if the pH changes more than about 0.2 units over short periods of time, automatic
pH correction is best. If the pH is relatively steady or subject only to seasonal changes, manual pH correction
is adequate.
5. INPUT FILTER. Before converting the sensor current to a chlorine reading, the analyzer applies an input filter.
The filter reduces noisy readings, but increases the response time. The level of filtering is selected by choosing the amount of time required for the display to reach 63% of a step change.
6. DUAL SLOPE CALIBRATION. The free chlorine sensor loses sensitivity at high concentrations of chlorine. The
FCL analyzer has a dual slope feature that allows the user to compensate for the non-linearity of the sensor.
For the vast majority of applications, dual slope calibration is unnecessary.
20
MODEL FCL
SECTION 5.0
PROGRAMMING THE ANALYZER
5.5.3 Definitions — pH/ORP
1. ORP. ORP is oxidation-reduction potential. It is the voltage difference between a noble metal (usually platinum) indicator electrode and a silver/silver chloride reference electrode. Not used with Model FCL.
2. REDOX. Redox is redox potential. Redox potential is measured the same way as ORP. The sign of the redox
potential is the negative of ORP. Not used with Model FCL.
3. GLASS IMPEDANCE FAULT. The analyzer continuously measures the impedance of the pH sensor glass
membrane. When the analyzer detects low glass impedance, indicating a broken or cracked glass membrane,
it automatically displays a fault message.
4. SOLUTION TEMPERATURE CORRECTION. The pH of a solution, particularly an alkaline one, is a function
of temperature. If the temperature changes, so does the pH, even though the concentration of the acid or base
causing the pH remains constant. Solution temperature compensation converts the pH at the measurement
temperature to the pH at a reference temperature (25°C).
NOTE
IN THE MODEL FCL, pH IS MEASURED FOR THE SOLE PURPOSE OF CORRECTING A FREE CHLORINE MEASUREMENT. DO NOT USE SOLUTION
TEMPERATURE CORRECTION. FREE CHLORINE READINGS MUST BE
CORRECTED USING THE pH AT THE ACTUAL SAMPLE TEMPERATURE.
5. ISOPOTENTIAL pH. Does not apply when pH is being measured to correct free chlorine readings.
21
MODEL FCL
SECTION 5.0
PROGRAMMING THE ANALYZER
5.5.4 Procedure.
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
Calibrate
Hold
Program
Display
Outputs
Alarms
Measurement
2. Choose Measurement.
>>
Configure?
Sensor1
Sensor2
S1 Chlorine Type
free
1. Press MENU. The main menu screen appears. Choose Program.
total
4. For Sensor 1 (chlorine), choose free.
>>
S1 pH Comp?
Auto
3. Choose Sensor 1 (chlorine) or Sensor 2 (pH). For a single input configuration (Model FCL-01), the Sensor 1 Sensor 2 screen does not appear.
If you chose Sensor 1, go to step 4.
If you chose Sensor 2, go to step 10.
Manual
Manual pH
5. Select Auto or Manual pH correction. For Model FCL-01, choose
Manual. For Model FCL-02, choose Auto.
6. If you chose Manual, enter the average pH of the sample. The analyzer will
use this pH in all subsequent calculations, no matter what the true pH is.
07.00 pH
Input filter?
7. Choose the amount of filtering desired.
63% in 005sec
Dual Range Cal?
Disable
Enable
8. Enable or disable dual slope calibration. In the vast majority of applications, dual slope calibration is unnecessary.
9. The display returns to the screen shown in step 3. To configure the pH
sensor, choose Sensor 2. To return to the previous screen, press EXIT.
To return to the main display, press MENU followed by EXIT.
pH
S2Measure?
Redox
11. Choose Yes.
S1 Glass Fault?
Enable?
10. If Sensor 2 (pH) was selected, the screen at left appears. Select pH.
ORP
Yes
No
12. Leave Soln Temp Corr and Sensor Isoptntl at their default values.
Soln Temp Corr
Sensor Isoptntl
a. Leave Soln Temp Corr turned Off.
S1 SolnTempCorr?
Off
Ultrapure
Sensor Isoptntl
S1:
>>
b. Leave Sensor Isoptntl at 7.00 pH.
07.00pH
13. The display returns to the screen shown in step 3. Press EXIT to return
to the previous screen. To return to the main display, press MENU followed by EXIT.
22
MODEL FCL
SECTION 5.0
PROGRAMMING THE ANALYZER
5.6 CHOOSING TEMPERATURE UNITS AND MANUAL OR AUTOMATIC TEMPERATURE
COMPENSATION
5.6.1 Purpose
This section describes how to do the following:
1. Choose temperature display units (°C or °F).
2. Choose automatic or manual temperature compensation for membrane permeability.
3. Choose automatic or manual temperature compensation for pH.
4. Enter a temperature for manual temperature compensation.
5.6.3 Definitions — chlorine
1. AUTOMATIC TEMPERATURE COMPENSATION - CHLORINE.
The chlorine sensor is a membrane-covered amperometric sensor. The permeability of the membrane is a
function of temperature. As temperature increases, membrane permeability increases. Thus, an increase in
temperature will cause the sensor current and the analyzer reading to increase even though the chlorine level
remained constant. A correction equation in the analyzer software automatically corrects for changes in membrane permeability caused by temperature. In automatic temperature compensation, the analyzer uses the
temperature measured by the sensor for the correction. Temperature is also used in the pH correction applied
to free chlorine measurements.
2. MANUAL TEMPERATURE COMPENSATION - CHLORINE. In manual temperature compensation, the analyzer uses the temperature entered by the user for membrane permeability and pH correction. It does not use
the actual process temperature. Do NOT use manual temperature compensation unless the measurement and
calibration temperatures differ by no more than about 2°C. Manual temperature compensation is useful if the
sensor temperature element has failed an a replacement sensor is not available.
5.6.3 Definitions — pH
1. AUTOMATIC TEMPERATURE COMPENSATION — pH. The analyzer uses a temperature-dependent factor
to convert measured cell voltage to pH. In automatic temperature compensation, the analyzer measures the
temperature and automatically calculates the correct conversion factor. For maximum accuracy, use automatic temperature compensation.
2. MANUAL TEMPERATURE COMPENSATION — pH. In manual temperature compensation, the analyzer converts measured voltage to pH using the temperature entered by the user. It does not use the actual process
temperature. Do NOT use manual temperature compensation unless the process temperature varies no more
than about ±2°C or the pH is between 6 and 8. Manual temperature compensation is useful if the sensor temperature element has failed and a replacement sensor is not available.
23
MODEL FCL
SECTION 5.0
PROGRAMMING THE ANALYZER
5.6.3 Procedure.
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
Calibrate
Hold
Program
Display
Outputs
Alarms
Measurement
2. Choose >>.
>>
Temp
Security
#Sensors
3. Choose Temp.
>>
Config Temp?
°C/F
1. Press MENU. The main menu screen appears. Choose Program.
Live/Manual
4. Choose °C/F to change temperature units. Choose Live/Manual to turn
on (Live) or turn off (Manual) automatic temperature compensation.
a. If °C/F is chosen, select °C or °F in the next screen.
b. If Live/Manual is chosen, select Live or Manual for sensor 1 (chlorine) in the next screen.
c. If Manual is chosen, enter the temperature in the next screen. The
temperature entered in this step will be used in all subsequent measurements, no matter what the process temperature is.
d. The display will return to the Live/Manual screen for sensor 2 (pH).
Make the desired selections for sensor 2.
5.7 SETTING A SECURITY CODE
5.7.1 Purpose.
This section describes how to set a security code. The security code prevents program and calibration settings
from accidentally being changed. Refer to Section 4.4 for additional information.
5.7.2 Procedure.
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
Calibrate
Hold
Program
Display
Outputs
Alarms
Measurement
Temp
#Sensors
>>
1. Press MENU. The main menu screen appears. Choose Program.
2. Choose >>, then Security.
Security
>>
3. Enter a three digit security code. The security code takes effect two minutes after the last key stroke.
4. The display returns to the security menu screen. Press EXIT to return to
the previous screen. To return to the main display, press MENU followed
by EXIT.
24
MODEL FCL
SECTION 5.0
PROGRAMMING THE ANALYZER
5.8 NOISE REJECTION
5.8.1 Purpose.
For maximum noise rejection, the frequency of the ac power must be entered in the analyzer.
5.8.2. Procedure.
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
Calibrate
Hold
Program
Display
Outputs
Alarms
Measurement
2. Choose >>.
>>
Temp
Security
#Sensors
3. Choose >>.
>>
Noise Rejection
ResetAnalyzer
1. Press MENU. The main menu screen appears. Choose Program.
4. Choose Noise Rejection.
>>
5. Enter the mains frequency, 50 Hz or 60 Hz.
6. The display returns to the Noise Rejection screen. To return to the main
menu, press EXIT. To return to the main display, press MENU followed by
EXIT.
5.9 SINGLE SENSOR OR DUAL SENSOR INPUT
5.9.1 Purpose
The FCL analyzer accepts input from a single chlorine sensor or from a chlorine and pH sensor. This section
describes how to program the analyzer for single or dual sensors. COMPLETE THIS SECTION BEFORE DOING
OTHER PROGRAMMING.
5.9.2 Procedure.
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
Calibrate
Hold
Program
Display
Outputs
Alarms
Measurement
Temp
#Sensors
2. Choose >>.
>>
Security
3. Choose #Sensors.
>>
# of sensors?
One
1. Press MENU. The main menu screen appears. Choose Program.
Two
4. Choose One for the Model FCL-01 (chlorine sensor only). Choose Two
for the Model FCL-02 (chlorine and pH sensor).
5. The display returns to the # Sensors screen. To return to the main menu,
press MENU. To return to the main display, press MENU followed by
EXIT.
25
MODEL FCL
SECTION 5.0
PROGRAMMING THE ANALYZER
5.10 RESETTING FACTORY CALIBRATION AND FACTORY DEFAULT SETTINGS
5.10.1 Purpose.
This section describes how to re-install factory calibration and default values. The process also clears all fault messages and returns the display to the first quick start screen.
5.10.2. Procedure.
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
Calibrate
Hold
Program
Display
Outputs
Alarms
Measurement
2. Choose >>.
>>
Temp
Security
#Sensors
3. Choose >>.
>>
Noise Rejection
ResetAnalyzer
4. Choose ResetAnalyzer.
>>
Load factory
settings?
1. Press MENU. The main menu screen appears. Choose Program.
Yes
No
5. Choose Yes or No. If Yes is selected, previous settings are cleared and
the Quick Start Menu appears.
5.11 SELECTING A DEFAULT SCREEN, LANGUAGE, AND SCREEN CONTRAST
5.11.1 Purpose
This section describes how to do the following:
1. set a default display screen
The default display screen is the screen shown during normal operation. The Model FCL analyzer allows the
user to choose from a number of screens. Which screens are available depends on how the analyzer was configured. The following is an explanation of the abbreviations used in the screens.
In the display:
i.
The units attached to the reading make clear what is being displayed, chlorine or pH. The units for
chlorine are ppm (mg/L) as Cl2.
ii.
S2 means sensor 2. S2 appears if the user has chosen to display data only from sensor 2 (pH).
iii. G is the impedance of the glass pH electrode.
2. select a language
3. change the screen contrast
26
MODEL FCL
SECTION 5.0
PROGRAMMING THE ANALYZER
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
5.11.2 Procedure: Selecting a Display Screen
Calibrate
Hold
Program
1. Press MENU. The main menu screen appears. Choose Display.
Display
2. Choose Default Display.
Default Display
Language
Contrst
3. Press " or ' until the desired display appears. Press ENTER. For an
explanation of abbreviations, see Section 5.11.1.
4. The display returns to the screen in step 2. To return to the main menu,
press MENU. To return to the main display, press MENU followed by
EXIT.
5.11.3 Procedure: Choosing a Language
Calibrate
Hold
Program
1. Press MENU. The main menu screen appears. Choose Display.
Display
2. Choose Language.
Default Display
Language
Contrast
English
Français
Español
>>
3. Choose English, Français, Español, Deutsch, Italiano, or Portugues.
4. The display returns to the screen in step 2. To return to the main menu,
press MENU. To return to the main display, press MENU followed by
EXIT.
5.11.2 Procedure: Changing Screen Contrast
Calibrate
Hold
Program
1. Press MENU. The main menu screen appears. Choose Display.
Display
2. Choose Contrst.
Default Display
Units
Contrst
Screen Contrast:
50
3. Press " or ' to increase or decrease the screen contrast. As contrast
increases, the number increases.
4. The display returns to the screen shown in step 2. To return to the main
menu, press MENU. To return to the main display, press MENU followed
by EXIT.
27
MODEL FCL
SECTION 6.0
CALIBRATION
SECTION 6.0
CALIBRATION
6.1
6.2
6.3
6.4
6.5
6.6
6.7
INTRODUCTION
CALIBRATING TEMPERATURE
CALIBRATION - FREE CHLORINE
AUTO CALIBRATION - pH
MANUAL CALIBRATION - pH
STANDARDIZATION - pH
ENTERING A KNOWN SLOPE - pH
6.1 INTRODUCTION
The Calibrate Menu allows the user to calibrate sensor 1 (chlorine) and sensor 2 (pH). The temperature element
in each sensor can also be calibrated.
Chlorine sensors require periodic full-scale calibration. The purpose of the full-scale standard is to establish the
slope of the calibration curve. Because stable chlorine standards do not exist, the sensor must be calibrated
against a test run a grab sample of the process liquid. Several manufacturers offer portable test kits for this
purpose.
New chlorine sensors must be zeroed before being placed in service. Sensors should also be zeroed every time
the electrolyte solution is replaced. Zeroing involves placing the sensor in a chlorine-free sample until the sensor
current drops to its lowest stable value.
For pH sensors, two-point buffer calibration is standard. In auto calibration the analyzer calculates the pH of the
buffer from the nominal value entered by the user and does not accept calibration data until readings are stable.
In manual calibration the user enters buffer values and judges when readings are stable. The pH reading can also
be standardized, that is, forced to match the reading from a referee instrument. Finally, if the user knows the electrode slope (at 25°C), he can enter it directly.
28
MODEL FCL
SECTION 6.0
CALIBRATION
6.2 CALIBRATING TEMPERATURE
6.2.1 Purpose
Temperature is important in the measurement of chlorine and pH for different reasons.
The free chlorine sensor is a membrane-covered amperometric sensor. As the sensor operates, free chlorine diffuses through the membrane and is consumed at an electrode immediately behind the membrane. The reaction
produces a current that depends on the rate at which the free chlorine diffuses through the membrane. The diffusion rate, in turn, depends on the concentration of the analyte and how easily it passes through the membrane (the
membrane permeability). Because membrane permeability is a function of temperature, the sensor current will
change if either the concentration or temperature changes. To account for changes in sensor current caused by
temperature alone, the analyzer automatically applies a membrane permeability correction. The membrane permeability changes about 3%/°C at 25°C, so a 1°C error in temperature produces about a 3% error in the reading.
Temperature also plays a role in the pH correction applied to free chlorine readings.
Temperature is also important in pH measurements.
1. The analyzer uses a temperature dependent factor to convert measured cell voltage to pH. Normally, a slight
inaccuracy in the temperature reading is unimportant unless the pH reading is significantly different from 7.00.
Even then, the error is small. For example, at pH 12 and 25°C, a 1°C error produces a pH error less than ±0.02.
2. During auto calibration, the analyzer recognizes the buffer being used and calculates the actual pH of the
buffer at the measured temperature. Because the pH of most buffers changes only slightly with temperature,
reasonable errors in temperature do not produce large errors in the buffer pH. For example, a 1°C error causes at most an error of ±0.03 in the calculated buffer pH.
Without calibration the accuracy of the temperature measurement is about ±0.4°C. Calibrate the sensor/analyzer
unit if
1. ±0.4°C accuracy is not acceptable
2. the temperature measurement is suspected of being in error. Calibrate temperature by making the analyzer
reading match the temperature measured with a standard thermometer.
29
MODEL FCL
SECTION 6.0
CALIBRATION
6.2.2 Procedure
1. Remove the sensor from the process. Place it in an insulated container of water along with a calibrated thermometer. Submerge at least the bottom two inches of the sensor. Stir continuously.
2. Allow the sensor to reach thermal equilibrium. For some sensors, the time constant for a change in temperature is 5 min., so it may take as long as 30 min. for temperature equilibration.
3. If the sensor cannot be removed from the process, measure the temperature of a flowing sample taken from
a point as close to the sensor as possible. Let the sample continuously overflow an insulated container holding a calibrated thermometer.
4. Change the analyzer display to match the calibrated thermometer using the procedure below.
Calibrate
Program
Hold
b. Choose Sensor1 (chlorine) or Sensor2 (pH).
Calibrate?
Sensor1
Sensor2
c.
CalSensor1?
Measurement
a. Press MENU. The main menu screen appears. Choose Calibrate.
Display
Choose Temp.
Temp
Live
25.0°C
CalS1
+25.0°C
d. If the analyzer was programmed in Section 5.6 to use the actual process
temperature, the screen at left will appear. To calibrate the temperature,
change the number in the second line to match the temperature measured with the standard thermometer. Press ENTER. Go to step f.
If the calibration temperature is more than 2 or 3°C different from the live
reading, see Section 8.3.1 or Section 8.4.2.
If the analyzer was programmed to use a temperature entered by the
user, go to step e.
S1: +25.0°C
e. The screen at left will appear. Change the temperature to the desired
value, then press ENTER. The analyzer will use the temperature entered
in this step in all measurements and calculations, no matter what the true
temperature is.
CalSensor1?
f.
ManualTemp?
Measurement
Temp
g. Choose the other sensor and calibrate its temperature response.
Calibrate?
Sensor1
The screen at left will appear. Press EXIT.
Sensor2
h. To return to the main display, press MENU followed by EXIT.
30
MODEL FCL
SECTION 6.0
CALIBRATION
6.3 CALIBRATION — FREE CHLORINE
6.3.1 Purpose
As Figure 6-1 shows, a free chlorine sensor generates a current directly proportional to the concentration of free
chlorine in the sample. Calibrating the sensor requires exposing it to a solution containing no chlorine (zero standard) and to a solution containing a known amount of chlorine (full-scale standard).
The zero standard is necessary because chlorine sensors, even when no chlorine is in the sample, generate a
small current called the residual current. The analyzer compensates for the residual current by subtracting it from
the measured current before converting the result to a chlorine value. New sensors require zeroing before being
placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. Either of the following makes a good zero standard:
•
Deionized water containing about 500 ppm sodium chloride. Dissolve 0.5 grams (1/8 teaspoonful) of table
salt in 1 liter of water. DO NOT USE DEIONIZED WATER ALONE FOR ZEROING THE SENSOR. THE
CONDUCTIVITY OF THE ZERO WATER MUST BE GREATER THAN 50 μS/cm.
•
Tap water known to contain no chlorine. Expose tap water to bright sunlight for at least 24 hours.
The purpose of the full-scale standard is to establish the slope of the calibration curve. Because stable chlorine
standards do not exist, the sensor must be calibrated against a test run on a grab sample of the process liquid. Several manufacturers offer portable test kits for this purpose. Observe the following precautions when taking and testing the grab sample.
•
Take the grab sample from a point as close to the sensor as possible. Be sure that taking the sample does
not alter the flow of the sample to the sensor. It is best to install the sample tap just downstream from the
sensor.
•
Chlorine solutions are unstable. Run the test immediately after taking the sample. Try to calibrate the sensor when the chlorine concentration is at the upper end of the normal operating range.
Free chlorine measurements also require a pH correction. Free chlorine is the sum of hypochlorous acid (HOCl)
and hyprochlorite ion (OCl-). The relative amount of each depends on the pH. As pH increases, the concentration
of HOCl decreases and the concentration of OCl- increases. Because the sensor responds only to HOCl, a pH
correction is necessary to properly convert the sensor current into a free chlorine reading.
The analyzer can use either automatic or manual pH correction. In automatic pH correction, the analyzer continuously monitors the pH of the solution and corrects the free chlorine reading for changes in pH. In manual pH correction, the user must enter the pH of the solution. Generally, if the pH changes more than about 0.2 units over
short periods of time, automatic pH correction is best. If the pH is relatively steady or subject only to seasonal
changes, manual pH correction is adequate.
During calibration, the analyzer must know the pH of the solution. If the analyzer is using automatic pH correction,
the pH sensor (properly calibrated) must be in the process liquid before starting the calibration. If the analyzer is using manual pH correction, be sure to enter the pH value before starting the calibration.
The free chlorine sensor loses sensitivity at high concentrations of chlorine. The FCL analyzer has a dual slope
feature that allows the user to compensate for the non-linearity of the sensor. However, for the vast majority of
applications, dual slope calibration is unnecessary.
FIGURE 6-1. Sensor Current as a Function of Free Chlorine
Concentration
31
MODEL FCL
SECTION 6.0
CALIBRATION
6.3.2 Procedure — Zeroing the sensor.
1. Place the sensor in the zero standard. See Section 6.3.1 for suggested zero standards. Be sure no air bubbles are trapped against the membrane. The sensor current will drop rapidly at first and then gradually reach
a stable zero value. To monitor the sensor current, go to the main display and press ' until the sensor input
current is showing. Typical zero current for a free chlorine sensor is in the range -10 to +10 nA.
A new sensor or a sensor in which the electrolyte solution has been replaced may require several hours (occasionally as long as overnight) to reach a minimum zero current. DO NOT START THE ZERO ROUTINE UNTIL
THE SENSOR HAS BEEN IN ZERO SOLUTION FOR AT LEAST TWO HOURS.
Calibrate
Hold
Program
2. Press MENU. The main menu screen appears. Choose Calibrate.
Display
Calibrate?
Sensor1
Sensor2
3. Choose Sensor 1 (free chlorine). For a single sensor configuration, this
screen will not appear.
4. Choose Measurement.
CalSensor1?
Measurement
Temp
5. Choose Zero.
Cal S1?
InProcess
Zero
S1 Live
1.000ppm
Zeroing
Wait
S1 Live
0.000ppm
Sensor Zero Done
Sensor Zero Fail
6. The screen at left appears. The top line is the current chlorine reading
based on the previous calibration or, for a first time calibration, the default
sensitivity.
7. Once the reading is stable, the screen at left appears. Sensor zero is
complete and the analyzer has stored the zero current. The screen
remains until the operator presses MENU then EXIT to return to the main
display.
NOTE
Pressing ENTER during the zero step will cause the analyzer to use the present sensor current as the zero current. If the
sensor is zeroed before the current has reached a minimum
stable value, subsequent readings will be in error.
After zeroing, leave the sensor in the zero solution and verify that the sensor current is between -10 and +10 nA. To display the sensor current, go
the the main display and press 'until the input current is showing.
8. This screen appears if the zero current is extremely high. See Section 8.3
for troubleshooting. To repeat the zero step, press EXIT and choose
Zero.
Current Too High
Possible ZeroErr
Proceed?
32
Yes
No
9. This screen appears if the zero current is moderately high. To continue,
choose Yes. To repeat the zero step, choose No.
MODEL FCL
SECTION 6.0
CALIBRATION
6.3.3 Procedure — Calibrating the sensor (single slope)
NOTE
Single slope calibration is the commonly used calibration method for free chlorine. Dual slope calibration, described is section 6.3.4, is rarely needed.
1. Place the chlorine sensor in the chlorine flow cell. If automatic pH correction is being used, calibrate the pH
sensor (section 6.4) and place it in the pH flow cell. If manual pH correction is being used, measure the pH of
the sample and enter the value. See Section 5.5. Adjust the sample flow until water overflows the center tube
in the constant head cup.
2. Adjust the chlorine concentration until it is near the upper end of the control range. Wait until the analyzer reading is stable before starting the calibration.
Calibrate
Hold
Program
3. Press MENU. The main menu screen appears. Choose Calibrate.
Display
Calibrate?
Sensor1
Sensor2
4. Choose Sensor 1 (free chlorine). For a single sensor configuration, this
screen will not appear.
5. Choose Measurement.
CalSensor1?
Measurement
Temp
6. Choose InProcess.
Cal S1?
InProcess
Zero
Live
2.000ppm
Cal S1
2.000ppm
7. The screen shown at left appears. The top line is the current chlorine
reading based on the previous calibration.
Sample the process liquid. Make a note of the reading before taking the
sample. Immediately determine free chlorine. Note the analyzer reading
again. If the present reading (X) differs from the reading when the sample was taken (Y), calculate the value to enter (C) from the following formula:
C = (X/Y) (A)
where A is the concentration of chlorine in the grab sample.
Change the reading in the second line to match the results of the grab
sample test.
8. During calibration, the analyzer stores the measured current and calculates the sensitivity. Sensitivity is sensor current in nA divided by the concentration of chlorine. The sensitivity of a 499ACL-01 (free chlorine) sensor is 250-350 nA/ppm at 25°C and pH 8.
Possible Cal Err
Proceed?
Calibration
Error
Yes
No
9. This screen appears if the sensitivity is much higher or lower than expected. See Section 8.3. for troubleshooting. To repeat the calibration step,
press EXIT and choose InProcess.
10. This screen appears if the sensitivity is moderately higher or lower than
expected. To continue, choose Yes. To repeat the calibration, choose No.
For troubleshooting assistance, see Section 8.3.
33
MODEL FCL
SECTION 6.0
CALIBRATION
6.3.4 Procedure — Calibrating the sensor (dual slope)
Figure 6-2 shows the principle of dual slope calibration.
Between zero and concentration C1, the sensor response
is linear. When the concentration of chlorine becomes
greater than C1, the response is non-linear. In spite of the
non-linearity, the response can be approximated by a
straight line between point 1 and point 2.
Dual slope calibration is rarely needed. It is probably useful in fewer than 5% of applications.
1. Place the sensor in the zero standard. See Section
6.3.1. Be sure no air bubbles are trapped against the
membrane. The sensor current will drop rapidly at first
and then gradually reach a stable zero value. To monitor the sensor current, go to the main display and
press 'until the sensor input current is showing.
Typical zero current for a free chlorine sensor is
between -10 and +10 nA.
FIGURE 6-2. Dual Slope Calibration
A new sensor or a sensor in which the electrolyte solution has been replaced may require several hours (occasionally as long as overnight) to reach a minimum zero current. DO NOT START THE ZERO ROUTINE UNTIL
THE SENSOR HAS BEEN IN ZERO SOLUTION FOR AT LEAST TWO HOURS.
2. Be sure the analyzer has been configured for dual slope calibration. See Section 5.5.4.
Calibrate
Hold
Program
Calibrate?
Sensor1
Sensor2
CalSensor1?
Temp
6. Choose Zero.
Cal S1?
pt1
pt2
S1 Live
1.000ppm
Zeroing
Wait
S1 Live
0.000ppm
Sensor
Zero
Done
Cal S1?
Zero
4. Choose Sensor 1 (free chlorine) for a single sensor configuration, this
screen will not appear.
5. Choose Measurement.
Measurement
Zero
3. Press MENU. The main menu screen appears. Choose Calibrate.
Display
pt1
pt2
7. The screen at left appears. The top line is the current chlorine reading
based on the previous calibration or, for a first time calibration, the default
sensitivity.
8. Once the reading is stable, the screen at left appears. Sensor zero is
complete, and the analyzer has stored the zero current. The screen
remains until the operator presses EXIT to return to the screen in step 9.
If a "Sensor zero fail" or "Possible zero error" screen appears, refer to
Section 8.3 -Troubleshooting.
9. Place the chlorine sensor in the chlorine flow cell. If automatic pH correction is being used, calibrate the pH sensor (see Section 6.3) and place
it in the pH flow cell. If manual pH correction is being used, measure the
pH of the sample and enter the value (see section 5.5). Adjust the sample flow until the water overflows the center tube in the constant head
cup.
Adjust the concentration of chlorine until it is near the upper end of the linear response range of the sensor. (pt1 in Figure 6-2).
34
MODEL FCL
SECTION 6.0
CALIBRATION
S1 Live
10.00ppm
pt1
10.00ppm
10. Choose pt1. The screen at left appears. The top line is the current chlorine reading based on the previous calibration or, for a first time calibration, the default sensitivity
11. Wait until the reading is stable.
Sample the process liquid. Make a note of the reading before taking the
sample. Immediately determine free chlorine. Note the analyzer reading
again. If the present reading (X) differs from the reading when the sample was taken (Y), calculate the value to enter (C) from the following formula:
C = (X/Y) (A)
where A is the concentration of chlorine in the grab sample.
Change the reading in the second line to match the results of the grab
sample test.
Cal S1?
Zero
pt1
pt2
12. The screen returns to the display in step 9.
13. Adjust the concentration of chlorine until it is near the top end of the range
(pt2 in Figure 6-2).
S1 Live
10.00ppm
pt2
10.00ppm
14. Choose pt2. The screen at left appears. The top line is the current chlorine reading based on the previous calibration or, for a first time calibration, the default sensitivity.
15. Following the procedure in step 11, determine chlorine in a sample of the
process liquid. Change the reading in the second line to match the results
of the grab sample test.
16. The display returns to the screen in step 9. Press MENU followed by EXIT
to return to the main display.
35
MODEL FCL
SECTION 6.0
CALIBRATION
6.4 AUTO CALIBRATION — pH
6.4.1 Purpose
1. New sensors must be calibrated before use. Regular recalibration is also necessary.
2. Use auto calibration instead of manual calibration. Auto calibration avoids common pitfalls and reduces errors.
6.4.2 Definitions
1. AUTO CALIBRATION. The analyzer recognizes the buffers and uses temperature-corrected pH values in the
calibration. The table lists the buffers the FCL analyzer recognizes.
pH at 25°C
(nominal pH)
Standard(s)
1.68
NIST, DIN 19266, JSI 8802, BSI (see note 1)
3.56
NIST, BSI
3.78
NIST
4.01
NIST, DIN 19266, JSI 8802, BSI
6.86
NIST, DIN 19266, JSI 8802, BSI
7.00
(see note 2)
7.41
NIST
9.18
NIST, DIN 19266, JSI 8802, BSI
10.01
NIST, JSI 8802, BSI
12.45
NIST, DIN 19266
Note 1: NIST is National Institute of Standards, DIN is Deutsche Institute für
Normung, JSI is Japan Standards Institute, and BSI is British Standards
Institute.
Note 2: pH 7 buffer is not a standard buffer. It is a popular commercial buffer
in the United States.
The analyzer also measures noise and drift and does not accept calibration data until readings are stable.
Calibration data will be accepted as soon as the pH reading is constant to within 0.02 units for 10 seconds.
The stability settings cannot be changed.
2. SLOPE AND OFFSET. Once the analyzer successfully completes the calibration, it calculates and displays
the calibration slope and offset. The slope is reported
as the slope at 25°C. Figure 6-3 defines the terms.
6.4.3 Procedure
1. Obtain two buffer solutions. Ideally the buffer pH values should bracket the range of pH values to be measured.
2. Remove the sensor from the flow cell. If the process
and buffer temperatures are appreciably different,
place the sensor in a container of tap water at the
buffer temperature. Do not start the calibration until the
sensor has reached the buffer temperature. Thirty minutes is usually adequate.
FIGURE 6-3. Calibration Slope and Offset
3. Calibrate the sensor by using the procedure on the following page.
36
MODEL FCL
Calibrate
Program
SECTION 6.0
CALIBRATION
Hold
Calibrate?
Sensor1
b. Choose Sensor2 (pH sensor).
Sensor2
c.
CalSensor2?
Measurement
S2
Slope
Standardize
d. Choose BufferCal.
BufferCal
e. Choose Auto.
Manual
f.
S2AutoCal?
Buffer1
Choose Measurement.
Temp
S2BufferCal?
Auto
a. Press MENU. The main menu screen appears. Choose Calibrate.
Display
Choose Buffer1.
Buffer2
g. Rinse the sensor with water and place it in buffer 1. Be sure the glass bulb
and reference junction are completely submerged. Swirl the sensor.
S2Live
AutoBuf1
7.00pH
Wait
S2Live
7.00pH
AutoBuf1
7.01pH
S2AutoCal?
Buffer1
S2Live
Buf2
7.00pH
Wait
7.00pH
AutoBuf2
7.01pH
Slope
Calibration
Error
The top line shows the actual reading (S2Live). The analyzer also identifies
the buffer and displays the nominal buffer value (buffer pH at 25°C). If the
displayed value is not correct, press p or q to select the correct value. The
nominal value will change, for example, from 7.01 pH to 6.86 pH. Press
ENTER.
j.
The screen shown at left appears.
k.
Remove the sensor from buffer 1, rinse it with water, and place it in buffer
2. Swirl the sensor. Choose Buffer2.
l.
The screen at left is displayed until the reading is stable (<0.02 pH change
in 10 sec). When the reading is stable, the screen in step m appears. To
bypass automatic stabilization, press ENTER at any time.
Buffer2
S2Live
S2Offset
i.
Buffer2
S2AutoCal?
Buffer1
h. The screen at left is displayed until the reading is stable (<0.02 pH change
in 10 sec). When the reading is stable, the screen in step i appears. To
bypass automatic stabilization, press ENTER at any time.
6mV
59.16@25°C
m. The top line shows the actual reading (S2Live). The analyzer also identifies
the buffer and displays the nominal buffer value (buffer pH at 25°C). If the
displayed value is not correct, press p or q to select the correct value. The
nominal value will change, for example, from 7.01 pH to 6.86 pH. Press
ENTER to accept the nominal value.
n. If the calibration was successful, the analyzer will display the offset and
slope (at 25°C). The display will return to the screen in step b. Choosing
Sensor1 (chlorine sensor) will permit the chlorine measurement to be calibrated.
o. If the slope is out of range (less than 45 mV/pH or greater than 60 mV/pH),
an error screen appears. The display then returns to step f. Repeat the calibration.
p. To return to the main display, press MENU followed by EXIT.
37
MODEL FCL
SECTION 6.0
CALIBRATION
6.5 MANUAL CALIBRATION — pH
6.5.1 Purpose
1. New sensors must be calibrated before use. Regular recalibration is also necessary.
2. Use manual calibration if non-standard buffers are being used; otherwise, use auto calibration. Auto calibration avoids common pitfalls and reduces errors.
6.5.2 Definitions
1. MANUAL CALIBRATION. In auto calibration the analyzer recognizes the buffer and uses the temperature-corrected pH value in the calibration. The analyzer also measures noise and drift and does not accept calibration
data until readings are stable. During manual calibration, the user must judge when readings are stable and
look up and enter the buffer values.
2. SLOPE AND OFFSET. Once the FCL successfully completes the calibration, it calculates and displays the calibration slope and offset. The slope is reported as the slope at 25ºC. Figure 6-1 defines the terms.
6.5.3 Procedure
1. Obtain two buffer solutions. Ideally, the buffer pH values should bracket the range of pH values to be measured. Also obtain a thermometer. The pH of most buffer solutions is a function of temperature. To calibrate the
sensor properly, the pH of the buffer at the measurement temperature must be entered in the analyzer.
2. Remove the sensor from the process liquid. If the process and buffer temperature are appreciably different,
place the sensor in a container of tap water at the buffer temperature. Do not start the calibration until the sensor has reached the buffer temperature. Thirty minutes is usually adequate.
3. Calibrate the sensor using the procedure on the following page.
38
MODEL FCL
SECTION 6.0
CALIBRATION
Calibrate
Program
Hold
Calibrate?
Sensor1
b. Choose Sensor2 (pH sensor).
Sensor2
c.
CalSensor2?
Measurement
S2
Slope
Choose Measurement.
Temp
Standardize
d. Choose BufferCal.
BufferCal
e. Choose Manual.
S2BufferCal?
Auto
Manual
f.
S2ManualCal?
Buffer1
a. Press MENU. The main menu screen appears. Choose Calibrate.
Display
Choose Buffer1.
Buffer2
g. Rinse the sensor with water and place it in buffer 1. Be sure the glass bulb
and junction are completely submerged. Swirl the sensor. Also place a
thermometer in the buffer. Press ENTER
S2Live
Buf1
7.00pH
07.00pH
S2ManualCal?
Buffer1
Buffer2
S2Live
10.00pH
Buf2
10.00pH
S2Offset
Slope
6mV
h. The top line shows the actual buffer reading (S2 Live). Wait until the reading is stable, then note the temperature. Change the pH in the second line
to the pH of the buffer at the measured temperature. Press ENTER.
i.
The screen at left appears. Choose Buffer2. Rinse the sensor and thermometer with water and place them in buffer 2. Be sure the bulb and junction are submerged. Swirl the sensor.
j.
The top line shows the actual buffer reading (S2 Live). Wait until the reading is stable, then note the temperature. Change the pH in the second line
to the pH of the buffer at the measured temperature. Press ENTER.
k.
If the calibration was successful, the analyzer will display the offset and
slope (at 25°C). The display will return to the screen in step b. Choosing
Sensor1 (chlorine sensor) will permit the chlorine measurement to be
calibrated.
l.
If the slope is out of range (less than 45 mV/pH or greater than 60 mV/pH),
an error screen appears. The display then returns to step f. Repeat the calibration.
59.16@25°C
Calibration Error!
m. To return to the main display, press MENU followed by EXIT.
39
MODEL FCL
SECTION 6.0
CALIBRATION
6.6 STANDARDIZATION — pH
6.6.1 Purpose
1. The pH measured by the FCL analyzer can be changed to match the reading from a second or referee instrument. The process of making the two readings agree is called standardization.
2. During standardization, the difference between the two pH values is converted to the equivalent voltage. The
voltage, called the reference offset, is added to all subsequent measured cell voltages before they are converted to pH. If a standardized sensor is placed in a buffer solution, the measured pH will differ from the buffer
pH by an amount equivalent to the standardization offset.
6.6.2 Procedure
1. Install the sensor in the flow cell.
2. Once readings are stable, measure the pH of the liquid using a referee instrument.
3. Because the pH of the process liquid may change if the temperature changes, measure the pH of the grab
sample immediately after taking it.
4. For poorly buffered samples, it is best to determine the pH of a continuously flowing sample from a point as
close as possible to the sensor.
5. Standardize the FCL analyzer by following the steps below.
Calibrate
Hold
Program
b. Choose Sensor2 (pH sensor).
Calibrate?
Sensor1
Sensor2
c.
CalSensor2?
Measurement
S2
Standardize
7.00pH
CalS2
07.00pH
40
e. The top line shows the present pH reading. Change the pH reading in the
second line to match the referee instrument. Press ENTER.
f.
Invalid Input!
Slope
d. Choose Standardize.
BufferCal
Live
S2
Choose Measurement.
Temp
Slope
Max:
a. Press MENU. The main menu screen appears. Choose Calibrate.
Display
14.00pH
Standardize
BufferCal
The screen at left appears if the entered pH was greater than 14.00. The
display then returns to step e. Repeat the standardization.
g. If the entry was accepted, the screen at left appears. To verify that the
new pH was accepted, return to the main display by pressing MENU followed by EXIT.
MODEL FCL
SECTION 6.0
CALIBRATION
6.7 ENTERING A KNOWN SLOPE VALUE — pH
6.7.1 Purpose
If the electrode slope is known from other measurements, it can be entered directly in the FCL analyzer. The slope
must be entered as the slope at 25°C. To calculate the slope at 25°C from the slope at temperature t°C, use the
equation:
298
slope at 25°C = (slope at t°C)
t°C + 273
Changing the slope overrides the slope determined from the previous buffer calibration.
6.7.2 Procedure
Calibrate
Hold
Program
Calibrate?
2. Choose Sensor2 (pH sensor).
Sensor1
Sensor2
3. Choose Measurement.
CalSensor2?
Measurement
S2
1. Press MENU. The main menu screen appears. Choose Calibrate.
Display
Temp
Standardize
Slope
4. Choose Slope.
BufferCal
5. The screen at left appears briefly.
Changing slope
overrides bufcal.
pH Slope @ 25°C?
S2:
59.16mV/pH
7. The slope must be between 45 and 60 mV/pH. If the value entered is outside this range, the screen at left appears.
Invalid Input!
S2
Slope
6. Change the slope to the desired value. Press ENTER.
Standardize
8. If the entry was accepted, the screen at left appears.
BufferCal
9. To return to the main display, press MENU followed by EXIT.
41
MODEL FCL
SECTION 7.0
MAINTENANCE
SECTION 7.0
MAINTENANCE
7.1
7.2
7.3
7.4
ANALYZER
CHLORINE SENSOR
pH SENSOR
CONSTANT HEAD SAMPLER
7.1 ANALYZER
The Model FCL analyzer needs little routine maintenance.
Clean the analyzer case and front panel by wiping with a clean soft cloth dampened with water ONLY. Do not
use solvent, like alcohol, that might cause a buildup of static charge.
Only a few components of the analyzer are replaceable. See Table 7-1 and Figure 7-1.
Circuit boards are not replaceable.
If the analyzer must be replaced, order as shown below.
Model
Order analyzer
FCL-01 (free chlorine only)
1055-01-11-24-68
FCL-02 (free chlorine and pH)
1055-01-11-24-32-68
Replacing the analyzer.
1. Turn off power to the FCL.
2. Loosen the four screws holding the front panel to the enclosure case and let the panel swing down.
3. Disconnect the power, alarm, output, and sensor wires.
4. The front panel is held to the rear enclosure by a
-shaped metal pin inserted into clips on each side
of the rear enclosure. To remove the front panel, close the panel until the pin moves slightly past the open end
of one of the clips. Using a small screwdriver, press down on the top of the clip. At the same time lift the pin
over the end of the clip. Once one end of the pin is free, the other end easily slides out of the other clip.
5. Using the procedure in step 4, remove the front panel from the replacement analyzer.
6. To install the replacement panel, place one end of the pin in one of the clips. Push the other end of the pin
over the other clip. The pin will snap into place.
7. Replace the power, alarm, output, and sensor wires. See Section 3.0 for wiring connections. For Model FCL-02
(free chlorine and pH) be sure to connect the jumper between TB5-6 and TB3-1. Be sure to leave adequate
slack to avoid stress on the conductors when the panel is opened.
42
MODEL FCL
SECTION 7.0
MAINTENANCE
TABLE 7-1. Replacement Parts for FCL Analyzer (1055-01-11-24-68 or 1055-01-11-24-32-68)
Location in
Figure 7-1
PN
Description
Shipping
Weight
1
note
Screw, 6-32 x 1.38 in.
2
note
O-ring 2-007
3
33655-00
Gasket for pipe/surface mount version
2 lb/1.0 kg
not shown
23833-00
Surface mount kit; consists of four self-tapping
screws #6 x 1.75 in. and four O-rings
1 lb/0.5 kg
Note: Information about the size of screws and O-rings is for information only. Screws and washers cannot be purchased from
Rosemount Analytical.
Shipping weights are rounded up to the nearest whole lb or 0.5 kg.
FIGURE 7-1. Exploded View of Model FCL Analyzer
43
MODEL FCL
SECTION 7.0
MAINTENANCE
7.2 CHLORINE SENSOR
CAUTION:
Fill solution may cause irritation.
May be harmful if swallowed.
Read and follow manual.
CAUTION:
PRESSURIZED SPRAY INJURY
Before removing the sensor from the process stream for
maintenance, be sure the process pressure is reduced to
0 psig and the process temperature is at a safe level!
7.2.1 General.
When used in clean water, the chlorine sensor requires little maintenance. Generally, the sensor needs maintenance when the response becomes sluggish or noisy or when readings drift following calibration. For a sensor
used in potable water, expect to clean the membrane every month and replace the membrane and electrolyte solution every three months. In water containing large amounts of suspended solids, for example open recirculating
cooling water, membrane cleaning or replacement will be more frequent.
7.2.2 Cleaning the membrane.
Clean the membrane with water sprayed from a wash bottle. Do not use tissues to clean the membrane.
7.2.3 Replacing the electrolyte solution and membrane.
1. Unscrew the membrane retainer and remove the membrane assembly and O-ring. See Figure 7-2.
2. Hold the sensor over a container with the cathode pointing down.
3. Remove the fill plug and allow the electrolyte solution to drain out.
4. Inspect the cathode. If it is tarnished, clean it using a cotton-tipped swab dipped in baking soda or alumina.
Use type A dry powder alumina intended for metallographic polishing of medium and soft metals. Rinse thoroughly with water.
5. Wrap the plug with several turns of pipe tape and set aside.
6. Prepare a new membrane. Hold the membrane assembly with the cup formed by the membrane and membrane holder pointing up. Fill the cup with electrolyte solution and allow the wooden ring to soak up the solution (usually takes several minutes).
7. Hold the sensor at about a 45-degree angle with the cathode end pointing up. Add electrolyte solution through
the fill hole until the liquid overflows. Tap the sensor near the threads to release trapped air bubbles. Add more
electrolyte solution if necessary.
8. Place the fill plug in the electrolyte port and begin screwing it in. After several threads have engaged, rotate
the sensor so that the cathode is pointing up and continue tightening the fill plug. Do not overtighten.
9. Place a new O-ring in the groove around the cathode post. Cover the holes at the base of the cathode stem
with several drops of electrolyte solution.
10. Insert a small blunt probe, like a toothpick with the end cut off, through the pressure equalizing port. See
Figure 7-2.
NOTE
Do not use a sharp probe. It will puncture the bladder and destroy the sensor.
Gently press the probe against the bladder several times to force liquid through the holes at the base of the
cathode stem. Keep pressing the bladder until no air bubbles can be seen leaving the holes. Be sure the holes
remain covered with electrolyte solution.
11. Place a drop of electrolyte solution on the cathode, then place the membrane assembly over the cathode.
Screw the membrane retainer in place.
12. The sensor may require several hours operating at the polarizing voltage to equilibrate after the electrolyte
solution has been replenished.
44
MODEL FCL
SECTION 7.0
MAINTENANCE
FIGURE 7-2. Chlorine Sensor Parts
TABLE 7-2. Spare Parts
33523-00
9550094
33521-00
23501-08
23502-08
9210356
Electrolyte Fill Plug
O-Ring, Viton 2-014
Membrane Retainer
Free Chlorine Membrane Assembly: includes one membrane assembly and one O-ring
Free Chlorine Membrane Kit: includes 3 membrane assemblies and 3 O-rings
#4 Free Chlorine Sensor Fill Solution, 4 oz (120 mL)
7.3 pH SENSOR
7.3.1 General.
When used in clean water, the pH sensor requires little maintenance. Generally, the sensor needs maintenance
when the response becomes sluggish or noisy. In clean water the typical cleaning frequency is once a month. In
water containing large amounts of suspended solids, for example open recirculating cooling water, cleaning frequency will be substantially greater.
7.3.2 Cleaning the Sensor
Remove soft deposits by rinsing with a stream of water from a wash bottle. If the sensor becomes coated with
rust, dissolve the rust by soaking the sensor in dilute hydrochloric acid (mix 5 mL of concentrated hydrochloric
acid with 100 mL of water) for no longer than 5 minutes at room temperature. Rinse the sensor thoroughly with
water and soak in pH 4 buffer for several hours. Recalibrate the sensor in buffers before returning it to service.
7.3.3 Other Maintenance
The 399VP-09-305 pH sensor supplied with the Model FCL-02 is disposable. It has no replaceable parts.
45
MODEL FCL
SECTION 7.0
MAINTENANCE
7.4 CONSTANT HEAD FLOW CONTROLLER
7.4.2 Cleaning the flow controller
7.4.1 General
The low flow controller can be taken apart completely
for cleaning. Use a strong flow of water to flush out the
tubing. A pipe cleaner or a small bottlebrush can
remove more adherent deposits. To prevent leaks,
apply a thin layer of silicone grease (or equivalent) to
the two O-rings at the base of overflow chamber and to
the O-ring sealing the central overflow tube to the base.
After a period of time, deposits may accumulate in the
constant head overflow chamber and in the tubing
leading to the flow cell(s). Deposits increase the resistance to flow and cause the flow to gradually decrease.
Loss of flow may ultimately have an impact on the chlorine sensor performance. The flow controller is
designed to provide about 2 gal/hr flow. Loss of flow to
about 1 gal/hr causes a 10-15% drop in chlorine sensor output. Loss of flow has almost no effect on pH
sensor performance other than to increase the overall
response time of the sensor.
46
7.4.3 Other Maintenance
Table 7-3 and Figure 7-3 show the replacement parts
for the flow controller assembly used in Model FCL-01.
MODEL FCL
SECTION 7.0
MAINTENANCE
TABLE 7-3. Replacement parts for constant head flow controller assembly (Model FCL-01)
Location in
Figure 7-3
PN
Description
Shipping
Weight
1
24091-01
Flow cell for chlorine sensor with bubble shedding nozzle
1 lb/0.5 kg
2
24040-00
O-ring kit, two 2-222 and one 2-024 silicone O-rings,
with lubricant
1 lb/0.5 kg
3
33812-00
Dust cap for constant head flow controller
1 lb/0.5 kg
4
9322032
Elbow, ¼ in FNPT x ¼ in OD tubing
1 lb/0.5 kg
5
9350029
Check valve, ¼ in FNPT
1 lb/0.5 kg
FIGURE 7-3. Replacement Parts for the Flow Controller Assembly used in Model FCL-01.
47
MODEL FCL
SECTION 7.0
MAINTENANCE
Table 7-4 and Figure 7-4 show the replacement parts for the flow controller assembly used in Model FCL-02.
FIGURE 7-4. Replacement Parts for the Flow Controller Assembly used in Model FCL-02.
TABLE 7-4. Replacement parts for constant head flow controller assembly (Model FCL-02)
Location in
Figure 7-4
PN
1
48
Description
Shipping
Weight
24091-01
Flow cell for chlorine sensor with bubble shedding nozzle
1 lb/0.5 kg
2
24091-00
Flow cell for pH sensor
1 lb/0.5 kg
3
24040-00
O-ring kit, two 2-222 and one 2-024 silicone O-rings,
with lubricant
1 lb/0.5 kg
4
33812-00
Dust cap for constant head flow controller
1 lb/0.5 kg
5
9322032
Elbow, ¼ in FNPT x ¼ in OD tubing
1 lb/0.5 kg
6
9350029
Check valve, ¼ in FNPT
1 lb/0.5 kg
MODEL FCL
SECTION 8.0
TROUBLESHOOTING
SECTION 8.0
TROUBLESHOOTING
8.1 OVERVIEW
8.2 TROUBLESHOOTING USING FAULT CODES
8.3 TROUBLESHOOTING WHEN NO ERROR MESSAGE IS SHOWING —
FREE CHLORINE
8.4 TROUBLESHOOTING WHEN NO ERROR MESSAGE IS SHOWING — pH
8.5 TROUBLESHOOTING WHEN NO ERROR MESSAGE IS SHOWING — GENERAL
8.6 SIMULATING INPUTS — CHLORINE
8.7 SIMULATING INPUTS — pH
8.8 SIMULATING TEMPERATURE
8.9 MEASURING REFERENCE VOLTAGE — pH
8.1 OVERVIEW
The analyzer used with the Model FCL continuously monitors itself and the sensor(s) for faults. When the analyzer detects a fault, the word fault appears in the display alternately with the measurement. If alarm 3 was configured as a fault alarm, the alarm relay will energize. The outputs do not change during a fault condition. They
continue to reflect the measured chlorine, pH or temperature. Press p to display the fault codes.
NOTE
A large number of information screens are available to aid troubleshooting. The most useful of these are
raw sensor current and sensitivity and zero current at last calibration. For pH measurements (available with
Model FCL-02), sensor slope and offset and glass impedance are also available. To view the information
screens, go to the main display and press the q key.
8.2 TROUBLESHOOTING USING FAULT CODES
Fault Code
S1 Out of Range
S2 Out of Range
S2 Broken Glass
TC1 Open or TC2 Open
TC1 Shorted or TC2 Shorted
S1 or S2 Sense Line Open
EEPROM Failure
Explanation
Sensor current exceeds 210 μA (chlorine only)
Absolute value of measured voltage exceeds 2500 mV (pH/ORP only)
pH sensitive glass membrane is broken
RTD for sensor 1 or sensor 2 is open
RTD for sensor 1 or sensor 2 is shorted
RTD sense line for sensor 1 or sensor 2 is open
EEPROM failure
See Section
8.2.1
8.2.2
8.2.3
8.2.4
8.2.4
8.2.5
8.2.6
8.2.1 Chlorine Sensor Current Exceeds 210 μA
Excessive sensor current implies that the chlorine sensor is miswired or the sensor has failed.
8.2.2 Absolute Value of Measured Voltage from the pH Sensor Exceeds 2500 mV
The voltage of a pH cell is usually between 600 mV and -600 mV. Readings outside the range -2500 mV to 2500
mV usually indicate a problem with sensor wiring or analyzer electronics.
A. If the sensor cable has just been replaced, check the wiring connections. See Section 3.2.
B. Verify that the sensor is completely submerged in the sample.
49
MODEL FCL
SECTION 8.0
TROUBLESHOOTING
8.2.3 pH Sensitive Glass Membrane is Broken
The analyzer continuously measures the impedance between the reference electrode and the inside of the pHsensing electrode. If the glass membrane is intact, the impedance is normally between 10 MΩ and 1000 MΩ. If the
membrane is cracked or broken, the impedance drops below 10 MΩ. If the membrane is cracked or broken, the
sensor must be replaced.
8.2.4 RTD for Sensor 1 or Sensor 2 Open or Shorted.
There is an open or short in the sensor RTD or wiring.
A. If the sensor cable has just been replaced, check the wiring connections. See Section 3.2.
B. Disconnect the sensor from the lead wire. Connect an ohmmeter across the RTD IN and RTD RETURN pins
on the Variopol plug at the top of the sensor. Refer to Figure 8-1 or 8-2. The resistance should be about 110
Ω. If there is an open or short circuit, the sensor has failed and should be replaced. If the resistance is acceptable, attach the sensor the Variopol cable and disconnect the RTD IN and RTD RETURN leads at the analyzer. Refer to Figure 3-2 or Figure 3-3. Connect an ohmmeter across the leads and measure the resistance.
If the circuit is open or shorted, the failure is in the cable, and the cable must be replaced.
C. If there is no open or short, check the analyzer. See Section 8.8.
FIGURE 8-1. Pin Out Diagram for
Model 499ACL-01-VP Sensor
(top view of connector end of sensor)
FIGURE 8-2. Pin Out Diagram for
Model 399VP-09 Sensor
(top view of connector end of sensor)
8.2.5 RTD Sense Line for Sensor 1 or Sensor 2 is Open.
The analyzer measures temperature using a three-wire RTD. See Figure 8-4. The in and return leads connect the
RTD to the measuring circuit in the analyzer. A third wire, called the sense line, is connected to the return line. The
sense line allows the analyzer to correct for the resistance of the in and return leads and to correct for changes in
lead wire resistance caused by changes in the ambient temperature.
A. Verify that all wiring connections are secure.
B. The system can be operated with the sense line open. The measurement will be less accurate because the
analyzer can no longer correct for lead wire resistance and for changes in lead wire resistance with ambient
temperature. However, if the sensor is to be used at approximately constant temperature, the lead wire resistance error can be eliminated by calibrating the sensor at the measurement temperature. Errors caused by
changes in lead wire resistance with changes in ambient temperature cannot be eliminated.To make the error
message disappear, connect the RTD sense and return terminals with a jumper.
8.2.6 EEPROM Failure.
Call the factory at (800) 854-8257.
50
MODEL FCL
SECTION 8.0
TROUBLESHOOTING
8.3 TROUBLESHOOTING WHEN NO ERROR MESSAGE IS SHOWING —
FREE CHLORINE
Problem
Zero current was accepted, but the current is outside the range -10 to 10 nA
Error or warning message appears while zeroing the sensor (zero current is too high)
Zero current is unstable
Sensor can be calibrated, but the current is less than about 250 nA/ppm at 25°C and pH 8
Process readings are erratic
Readings drift
Sensor does not respond to changes in chlorine level
Chlorine reading spikes following rapid change in pH
See Section
8.3.1
8.3.1
8.3.2
8.3.3
8.3.4
8.3.5
8.3.6
8.3.7
8.3.1 Zero current is too high
A. Is the sensor properly wired to the analyzer? See Section 3.2.
B. Is the zero solution chlorine-free? Take a sample of the solution and test it for free chlorine level. The concentration should be less than 0.02 ppm.
C. Has adequate time been allowed for the sensor to reach a minimum stable residual current? It may take several hours, sometimes as long as overnight, for a new sensor to stabilize.
D. Check the membrane for damage and replace it if necessary.
8.3.2 Zero current is unstable
A. Is the sensor properly wired to the analyzer? See Section 3.2. Verify that all wiring connections are tight.
B. Readings are often erratic when a new or rebuilt sensor is first placed in service. Readings usually stabilize
after about an hour.
C. Is the conductivity of the zero solution greater than 50 µS/cm? DO NOT USE DEIONIZED OR DISTILLED
WATER TO ZERO THE SENSOR. The zero solution should contain at least 0.5 grams of sodium chloride per
liter.
D. Is the space between the membrane and cathode filled with electrolyte solution and is the flow path between
the electrolyte reservoir and membrane clear? Often the flow of electrolyte and be started by simply holding
the sensor with the membrane end pointing down and sharply shaking the sensor a few times as though shaking down a clinical thermometer.
If shaking does not work, try clearing the holes around the cathode stem. Hold the sensor with the membrane
end pointing up. Unscrew the membrane retainer and remove the membrane assembly. Be sure the wood ring
remains with the membrane assembly. Use the end of a straightened paper clip to clear the holes at the base
of the cathode stem. Replace the membrane.
Verify that the sensor is filled with electrolyte solution. Refer to Section 7.2.
51
MODEL FCL
SECTION 8.0
TROUBLESHOOTING
8.3.3 Sensor can be calibrated, but the current is too low
A. Is the temperature low or is the pH high? Sensor current is a strong function of pH and temperature. The sensor current decreases about 3% for every °C drop in temperature. Sensor current also decreases as pH
increases. Above pH 7, a 0.1 unit increase in pH lowers the current about 5%.
B. Sensor current depends on the rate of sample flow past the sensor tip. If the flow is too low, chlorine readings
will be low. Verify that the chlorine sensor is installed in the correct flow cell. See Figure 2-1. Be sure the liquid level in the constant head sampler is level with the central overflow tube and that excess sample is flowing down the tube. If necessary, disassemble and clean the overflow sampler. See Section 7.4.
C. Low current can be caused by lack of electrolyte flow to the cathode and membrane. See step D in Section
8.3.2.
D. Is the membrane fouled or coated? A dirty membrane inhibits diffusion of free chlorine through the membrane,
reducing the sensor current and increasing the response time. Clean the membrane by rinsing it with a stream
of water from a wash bottle. DO NOT use a membrane or tissue to wipe the membrane.
E. If cleaning the membrane does not improve the sensor response, replace the membrane and electrolyte solution. If necessary, polish the cathode. See Section 7.2 for details.
8.3.4 Process readings are erratic
A. Readings are often erratic when a new sensor or a rebuilt sensor is first placed in service. The current usually stabilizes after a few hours.
B. Are the holes between the membrane and the electrolyte reservoir open? Refer to Section 8.3.2.
C. Verify that wiring is correct. Pay particular attention to shield and ground connections.
D. If automatic pH correction is being used, check the pH reading. If the pH reading is noisy, the chlorine reading will also be noisy. If the pH sensor is the cause of the noise, use manual pH correction until the problem
with the pH sensor can be corrected. Also, refer to Section 8.4.8 for troubleshooting noisy pH readings.
E. Is the membrane in good condition and is the sensor filled with electrolyte solution? Replace the fill solution
and electrolyte. Refer to Section 7.2 for details.
8.3.5 Readings drift
A. Is the sample temperature changing? Membrane permeability is a function of temperature. The time constant
for the 499ACL-01 sensor is about five minutes. Therefore, the reading may drift for a while after a sudden
temperature change.
B. Is the membrane clean? For the sensor to work properly, chlorine must diffuse freely through the membrane.
A coating on the membrane will interfere with the passage of chlorine, resulting in slow response. Clean the
membrane by rinsing it with a stream of water from a wash bottle. DO NOT use a membrane or tissue to wipe
the membrane.
C. Is the sample flow within the recommended range? Gradual loss of sample flow will cause a downward drift.
Be sure the liquid level in the constant head sampler is level with the central overflow tube and that excess
sample is flowing down the tube. If necessary, disassemble and clean the overflow sampler. See Section 7.4.
D. Is the sensor new or has it been recently serviced? New or rebuilt sensors may require several hours to stabilize.
E. Is the pH of the process changing? If manual pH correction is being used, a gradual change in pH will cause
a gradual change in the chlorine reading. As pH increases, chlorine readings will decrease, even though the
free chlorine level (as determined by a grab sample test) remained constant. If the pH change is no more than
about 0.2, the change in the chlorine reading will be no more than about 10% of reading. If the pH changes
are more than 0.2, use automatic pH correction.
52
MODEL FCL
SECTION 8.0
TROUBLESHOOTING
8.3.6 Sensor does not respond to changes in chlorine level.
A. Is the grab sample test accurate? Is the grab sample representative of the sample flowing to the sensor?
B. Is sample flowing past the sensor? Be sure the liquid level in the constant head sampler is level with the central overflow tube and that excess sample is flowing down the tube. If necessary, disassemble and clean the
overflow sampler. See Section 7.4.
C. Is the pH compensation correct? If the analyzer is using manual pH correction, verify that the pH value in the
analyzer equals the actual pH to within ±0.1 pH. If the analyzer is using automatic pH correction, check the
calibration of the pH sensor.
D. Is the membrane clean? Clean the membrane and replace it if necessary. Check that the holes at the base of
the cathode stem are open. Use a straightened paper clip to clear blockages. Replace the electrolyte solution.
E. Replace the sensor.
8.3.7 Chlorine readings spike following sudden changes in pH (automatic pH correction).
Changes in pH alter the relative amounts of hypochlorous acid (HOCl) and hypochlorite ion (OCl-) in the sample. Because the sensor responds only to HOCl, an increase in pH causes the sensor current (and the apparent chlorine level) to drop even though the actual free chlorine concentration remained constant. To correct for
the pH effect, the controller automatically applies a correction. Generally, the pH sensor responds faster than
the chlorine sensor. After a sudden pH change, the controller will temporarily over-compensate and gradually
return to the correct value. The time constant for return to normal is about 5 minutes.
8.3.8 Chlorine readings are too low.
A. Was the sample tested as soon as it was taken? Chlorine solutions are unstable. Test the sample immediately after collecting it. Avoid exposing the sample to sunlight.
B. Low readings can be caused by zeroing the sensor before the residual current has reached a stable minimum
value. Residual current is the current the sensor generates even when no chlorine is in the sample. Because
the residual current is subtracted from subsequent measured currents, zeroing before the current is a minimum can lead to low results.
Example: The true residual current for a free chlorine sensor is 4 nA, and the sensitivity is 350 nA/ppm.
Assume the measured current is 200 nA. The true concentration is (200-4)/350 or 0.56 ppm. If the sensor was
zeroed prematurely when the current was 10 nA, the measured concentration will be (200-10)/350 or 0.54
ppm. The error is 3.6%. Suppose the measured current is 400 nA. The true concentration is 1.13 ppm, and
the measured concentration is 1.11 ppm. The error is now 1.8%. The absolute difference between the reading
remains the same, 0.02 ppm.
C. Sensor response depends on flow. If the flow is too low, readings will be low and flow sensitive. Verify that the
chlorine sensor is installed in the correct flow cell. See Figure 2-1. Verify that the liquid level in the constant
head sampler is level with the central overflow tube and that excess sample is flowing down the tube. If necessary, disassemble and clean the overflow sampler. See Section 7.4.
53
MODEL FCL
SECTION 8.0
TROUBLESHOOTING
8.4 TROUBLESHOOTING WHEN NO ERROR MESSAGE IS SHOWING — pH.
Problem
See Section
New temperature during calibration more than 2-3°C different from the live reading
8.4.1
Calibration Error warning during two-point calibration
8.4.2
Calibration Error warning during standardization
8.4.3
Invalid Input while manually entering slope
8.4.4
Sensor does not respond to known pH changes
8.4.5
Calibration was successful, but process pH is slightly different from expected value
8.4.6
Calibration was successful, but process pH is grossly wrong and/or noisy
8.4.7
pH readings are moderately noisy and tend to wander
8.4.8
8.4.1 Difference Between Analyzer and Standard Thermometer is Greater Than 3°C.
A. Is the standard thermometer, RTD, or thermistor accurate? General purpose liquid-in-glass thermometers, particularly ones that have been mistreated, can have surprisingly large errors.
B. Is the temperature element in the pH sensor completely submerged in the test liquid?
C. Is the standard temperature sensor submerged to the correct level?
D. Review Section 6.2.
8.4.2 Calibration Error During Two-Point Calibration
Once the two-point (manual or automatic) calibration is complete, the analyzer automatically calculates the sensor slope (at 25°). If the slope is greater than 60 mV/pH or less than 45 mV/pH, the analyzer displays the
Calibration Error screen and does not update the calibration. Check the following:
A. Are the buffers accurate? Inspect the buffers for obvious signs of deterioration, such as turbidity or mold
growth. Neutral and slightly acidic buffers are highly susceptible to molds. Alkaline buffers (pH 9 and greater),
if they have been exposed to air for long periods, may also be inaccurate. Alkaline buffers absorb carbon dioxide from the atmosphere, which lowers the pH. If a high pH buffer was used in the failed calibration, repeat the
calibration using a fresh buffer. If fresh buffer is not available, use a lower pH buffer. For example, use pH 4
and 7 buffer instead of pH 7 and 10 buffer.
B. Was adequate time allowed for temperature equilibration? If the sensor was in a process substantially hotter
or colder than the buffer, place it in a container of water at ambient temperature for at least 20 minutes before
starting the calibration. Using auto calibration avoids calibration errors caused by temperature drift. The analyzer will not update readings until the drift is less than 0.02 pH over 10 seconds.
C. Were correct pH values entered during manual calibration? Using auto calibration eliminates errors caused by
improperly entering data.
D. Is the sensor properly wired to the analyzer? See Section 3.2.
E. Is the sensor dirty or coated? See Section 7.3.2..
Glass Imp
123M^
F.
Is the sensor faulty? With the main display showing, use p or q to scroll through
the information screens until the electrode impedance screen (at left) is displayed.
Refer to the table on the following page for an interpretation of the impedance
readings.
Another way of checking for a faulty sensor is to replace it with a new one. If the new sensor can be calibrated, the old sensor has failed.
G. Is the analyzer faulty? The best way to check for a faulty analyzer is to simulate pH inputs. See Section 8.9.
54
MODEL FCL
SECTION 8.0
TROUBLESHOOTING
GLASS IMPEDANCE (Glass Imp)
less than 10 MΩ
Glass bulb is cracked or broken. Sensor has failed.
between 10 and 1000 MΩ
Normal reading.
greater than 1000 MΩ
pH sensor may be nearing the end of its service life.
8.4.3 Calibration Error during Standardization.
During standardization, the millivolt signal from the pH cell is increased or decreased until it agrees with the pH
reading from a referee instrument. A unit change in pH requires an offset of about 59 mV. The analyzer limits the
offset to ±1400 mV. If the standardization causes an offset greater than ±1400 mV, the analyzer will display the
Calibration Error screen. The standardization will not be updated. Check the following:
A. Is the referee pH meter working and properly calibrated? Check the response of the referee sensor in buffers.
B. Is the process sensor working properly? Check the process sensor in buffers.
C. Is the sensor fully immersed in the process liquid? If the sensor is not completely submerged, it may be measuring the pH of the liquid film covering the glass bulb and reference element. The pH of this film may be different from the pH of the bulk liquid.
D. Is the sensor fouled? The sensor measures the pH of the liquid adjacent to the glass bulb. If the sensor is
heavily fouled, the pH of liquid trapped against the bulb may be different from the bulk liquid.
E. Has the sensor been exposed to poisoning agents (sulfides or cyanides) or has it been exposed to extreme
temperature? Poisoning agents and high temperature can shift the reference voltage many hundred millivolts.
To check the reference voltage, see Section 8.9.
8.4.4 Invalid Input While Manually Entering Slope.
If the sensor slope is known from other sources, it can be entered directly into the analyzer. The FCL will not accept
a slope (at 25°C) outside the range 45 to 60 mV/pH. See section 8.4.2 for troubleshooting sensor slope problems.
8.4.5 Sensor Does Not Respond to Known pH Changes.
A. Did the expected pH change really occur? If the process pH reading was not what was expected, check the
performance of the sensor in buffers. Also, use a second pH meter to verify the change.
B. Is sample flowing past the sensor? Be sure the liquid level in the constant head sampler is level with the central overflow tube and that excess sample is flowing down the tube. If necessary, disassemble and clean the
overflow sampler. See Section 7.4.
C. Is the sensor properly wired to the analyzer? See Section 3.2.
D. Is the glass bulb cracked or broken? Check the glass electrode impedance. See Section 8.4.2.
E. Is the analyzer working properly. Check the analyzer by simulating the pH input. See Section 8.7.
8.4.6 Buffer Calibration Is Acceptable, Process pH Is Slightly Different from Expected Value.
Differences between pH readings made with an on-line instrument and a laboratory or portable instrument are normal. The on-line instrument is subject to process variables, for example ground potentials, stray voltages, and orientation effects that may not affect the laboratory or portable instrument. To make the process reading agree with
a referee instrument, see Section 6.5.
55
MODEL FCL
SECTION 8.0
TROUBLESHOOTING
8.4.7 Calibration Was Successful, but Process pH Is Grossly Wrong and/or Noisy.
Grossly wrong or noisy readings suggest a ground loop (measurement system connected to earth ground at more
than one point), a floating system (no earth ground), or noise being brought into the analyzer by the sensor cable.
The problem arises from the process or installation. It is not a fault of the analyzer. The problem should disappear
once the sensor is taken out of the system. Check the following:
A. Is a ground loop present?
1. Verify that the system works properly in buffers. Be sure there is no direct electrical connection between
the buffer containers and the process liquid or piping.
2. Strip back the ends of a heavy gauge wire. Connect one end of the wire to the process piping or place it
in the process liquid. Place the other end of the wire in the container of buffer with the sensor. The wire
makes an electrical connection between the process and sensor.
3. If offsets and noise appear after making the connection, a ground loop exists.
B. Is the process grounded?
1. The measurement system needs one path to ground: through the process liquid and piping. Plastic piping, fiberglass tanks, and ungrounded or poorly grounded vessels do not provide a path. A floating system
can pick up stray voltages from other electrical equipment.
2. Ground the piping or tank to a local earth ground.
3. If noise still persists, simple grounding is not the problem. Noise is probably being carried into the instrument through the sensor wiring.
C. Simplify the sensor wiring.
1. Disconnect all sensor wires at the analyzer except pH/mV IN, REFERENCE IN, RTD IN and RTD
RETURN. See the wiring diagrams in Section 3.2.
2. Tape back the ends of the disconnected wires to keep them from making accidental connections with other
wires or terminals.
3. Connect a jumper wire between the RTD RETURN and RTD SENSE terminals (see wiring diagrams in
Section 3.2).
4. If noise and/or offsets disappear, the interference was coming into the analyzer through one of the sensor
wires. The system can be operated permanently with the simplified wiring.
D. Check for extra ground connections or induced noise.
1. To avoid induced noise in the sensor cable, keep the unit as far away as possible from power cables,
relays, and electric motors.
2. If ground loops persist, consult the factory. A visit from an experienced technician may be required to solve
the problem.
8.4.8 pH Readings Are Moderately Noisy and Tend to Wander.
pH readings that are moderately noisy (±0.1 pH) and tend to wander are probably caused by bubbles getting
trapped against the pH sensor. Although the overflow sampler is designed to allow bubbles to escape before they
reach the pH sensor and the sensor itself is designed so trapped air bubbles don’t interfere with the measurement,
bubbles may occasionally be a problem. Shaking the sensor will dislodge the bubbles. If bubbles remain a problem, contact the factory.
56
MODEL FCL
SECTION 8.0
TROUBLESHOOTING
8.5 TROUBLESHOOTING WHEN NO ERROR MESSAGE IS SHOWING — GENERAL
Problem
See Section
Current output is too low
8.5.1
Alarm relays do not operate when setpoint is exceeded
8.5.2
Display is unreadable — too faint or all pixels dark
8.5.3
8.5.1 Current Output Too Low.
Load resistance is too high. Maximum load is 600 Ω.
8.5.2 Alarm Relays Do Not Work
Verify the relays are properly wired.
8.5.3 Display is Unreadable.
While holding down the MENU key, press p or q until the display has the correct contrast.
8.6 SIMULATING INPUTS — CHLORINE
To check the performance of the analyzer, use a decade box and battery to simulate the current from the sensor. The
battery, which opposes the polarizing voltage, is necessary to ensure that the sensor current has the correct sign.
A. Disconnect the anode and cathode leads from terminals 1 and 2 on TB3 and connect a decade box and battery as shown in Figure 8-3. It is not necessary to disconnect the RTD leads.
B. Set the decade box to the resistance shown below.
Polarizing Voltage
200 mV
Resistance
Expected Current
28 MΩ
500 nA
C. Note the sensor current. It should be close to the value in the table. The actual value depends on the voltage
of the battery. To view the sensor current, go to the main display and press q until the sensor current is displayed.
D. Change the decade box resistance and verify that the correct current is shown. Calculate current from the
equation:
Vbattery — Vpolarizing (mV)
current (μA) =
resistance (kΩ)
The voltage of a fresh 1.5 volt battery is about 1.6 volt (1600 mV).
FIGURE 8-3. Simulating Chlorine
57
MODEL FCL
SECTION 8.0
TROUBLESHOOTING
8.7 SIMULATING INPUTS — pH
8.7.1 General
This section describes how to simulate a pH input into the Model FCL analyzer. To simulate a pH measurement,
connect a standard millivolt source to the analyzer. If the analyzer is working properly, it will accurately measure
the input voltage and convert it to pH.
8.7.2 Simulating pH input.
1. Turn off automatic temperature correction and solution temperature correction. From the Program menu, choose Temp. Then
choose Live/Manual and enter 25°C. See Section 5.6 for
details.
2. Disconnect the sensor and connect a jumper wire between the
pH IN and REFERENCE IN terminals.
3. From the display menu choose the pH/temperature/mV screen.
The measured voltage should be 0 mV and the pH should be
7.00. Because calibration data stored in the analyzer may be
offsetting the input voltage, the displayed pH may not be exactly 7.00.
FIGURE 8-4. Simulating pH Inputs
4. If a standard millivolt source is available, disconnect the jumper wire between pH IN and REFERENCE IN and
connect the voltage source as shown in Figure 8-4. Be sure to jumper the reference and solution ground terminals.
5. Calibrate the analyzer using the procedure in Section 6.3. Use 0.0 mV for Buffer 1 (pH 7.00) and -177.4 mV
for Buffer 2 (pH 10.00). If the analyzer is working properly it should accept the calibration. The slope should
be 59.16 mV/pH and the offset should be zero.
6. To check linearity, return to the main display and the pH/temperature/mV screen. Set the voltage source to the
values shown in the table and verify that the pH and millivolt readings match the values in the table.
Voltage (mV)
295.8
177.5
59.2
-59.2
-177.5
-295.8
58
pH (at 25°)
2.00
4.00
6.00
8.00
10.00
12.00
MODEL FCL
SECTION 8.0
TROUBLESHOOTING
8.8 SIMULATING TEMPERATURE
8.8.1 General.
The FCL analyzer accepts a Pt100 RTD (for pH and
chlorine sensors). The Pt100 RTD is in a three-wire
configuration. See Figure 8-5.
8.8.2 Simulating temperature
To simulate the temperature input, wire a decade box to
the analyzer or junction box as shown in Figure 8-6.
To check the accuracy of the temperature measurement, set the resistor simulating the RTD to the values
indicated in the table and note the temperature readings. The measured temperature might not agree with
the value in the table. During sensor calibration an offset
might have been applied to make the measured temperature agree with a standard thermometer. The offset
is also applied to the simulated resistance. The FCL is
measuring temperature correctly if the difference
between measured temperatures equals the difference
between the values in the table to within ±0.1°C.
FIGURE 8-5. Three-Wire RTD Configuration.
Although only two wires are required to connect
the RTD to the analyzer, using a third (and sometimes fourth) wire allows the analyzer to correct
for the resistance of the lead wires and for
changes in the lead wire resistance with temperature.
For example, start with a simulated resistance of 103.9
Ω, which corresponds to 10.0°C. Assume the offset from
the sensor calibration was -0.3 Ω. Because of the offset,
the analyzer calculates temperature using 103.6 Ω. The
result is 9.2°C. Now change the resistance to 107.8 Ω,
which corresponds to 20.0°C. The analyzer uses 107.5 Ω
to calculate the temperature, so the display reads
19.2°C. Because the difference between the displayed
temperatures (10.0°C) is the same as the difference
between the simulated temperatures, the analyzer is
working correctly.
FIGURE 8-6. Simulating RTD Inputs.
Temp. (°C)
0
10
20
25
30
40
50
60
70
80
85
90
100
Pt 100 (Ω)
100.0
103.9
107.8
109.7
111.7
115.5
119.4
123.2
127.1
130.9
132.8
134.7
138.5
59
MODEL FCL
SECTION 8.0
TROUBLESHOOTING
8.9 MEASURING REFERENCE VOLTAGE
Some processes contain substances that poison or
shift the potential of the reference electrode. Sulfide
is a good example. Prolonged exposure to sulfide
converts the reference electrode from a silver/silver
chloride electrode to a silver/silver sulfide electrode.
The change in reference voltage is several hundred
millivolts. A good way to check for poisoning is to
compare the voltage of the reference electrode with
a silver/silver chloride electrode known to be good.
The reference electrode from a new sensor is best.
See Figure 8-7. If the reference electrode is good,
the voltage difference should be no more than
about 20 mV. A poisoned reference electrode usually requires replacement.
FIGURE 8-7. Checking for a Poisoned Reference
Electrode.
Refer to the sensor wiring diagram to identify the reference
leads. A laboratory silver/silver chloride electrode can be
used in place of the second sensor.
60
MODEL FCL
SECTION 9.0
RETURN OF MATERIAL
SECTION 9.0
RETURN OF MATERIAL
9.1
9.2
9.3
GENERAL
WARRANTY REPAIR
NON-WARRANTY REPAIR
9.1 GENERAL.
To expedite the repair and return of instruments, proper communication between the customer and the factory is
important. Before returning a product for repair, call 1-949-757-8500 for a Return Materials Authorization (RMA)
number.
9.2 WARRANTY REPAIR.
The following is the procedure for returning instruments still under warranty:
1.
Call Rosemount Analytical for authorization.
2.
To verify warranty, supply the factory sales order number or the original purchase order number. In the case
of individual parts or sub-assemblies, the serial number on the unit must be supplied.
3.
Carefully package the materials and enclose your “Letter of Transmittal” (see Warranty). If possible, pack the
materials in the same manner as they were received.
4.
Send the package prepaid to:
Emerson Process Management
Liquid Division
2400 Barranca Parkway
Irvine, CA 92606
Attn: Factory Repair
RMA No. ____________
Mark the package: Returned for Repair
Model No. ____
9.3 NON-WARRANTY REPAIR.
The following is the procedure for returning for repair instruments that are no longer under warranty:
1.
Call Rosemount Analytical for authorization.
2.
Supply the purchase order number, and make sure to provide the name and telephone number of the individual to be contacted should additional information be needed.
3.
Do Steps 3 and 4 of Section 9.2.
NOTE
Consult the factory for additional information regarding service or repair.
61
The right people, the right answers, right now.
Immediate, Reliable Analytical Support
Now there’s a way to quickly get the right answers for your liquid analytical instrumentation questions: the Analytical
Customer Support Center.
Our staff of trained professionals is ready to provide the information you need. If you are placing an order, verifying delivery,
requesting application information, or just want to contact an Emerson Process Management representative, a call to the
Customer Support Center will provide you with the right people, the right answers, right now.
A Worldwide Network of Sales and Service
Emerson Process Management’s field sales offices are your source for more information on the fill line of Rosemount Analytical
products. Field sales personnel will work closely with you to supply technical data and application information.
For more information, please contact your nearest Emerson Process Management sales office.
THE AMERICAS HEADQUARTERS
Emerson Process Management
Rosemount Analytical Inc.
Liquid Center of Excellence
2400 Barranca Parkway
Irvine, CA 92606
Phone: +1.949.757.8500
Toll Free: +1.800.854.8257
Fax: +1.949.474.7250
ASIA-PACIFIC
Emerson Process Management
Asia Pacific Private Ltd.
1 Pandan Crescent
Singapore 0512
Republic of Singapore
Phone: 65.777.8211
Fax: 65.777.0947
GERMANY
Emerson Process Management Process
Gas Analyzer Center of Excellence
GmbH & Co. OHG
Industriestrasse 1
63594 Hasselroth
Germany
T 49.6055.884.0
F 49.6055.884.20
LATIN AMERICA
Emerson Process Management
Rosemount Analytical
10241 West Little York, Suite #200
Houston, TX 77040 USA
T 713.467.6000
F 713.827.3328
MIDDLE EAST AND AFRICA
EUROPE
Emerson Process Management AG
Blegistrasse 21
CH-6341 Baar-Walterswil
Switzerland
T 41.41.768.6111
T 41.41.761.8740
VISIT OUR WEBSITE AT
www.rosemountanalytical.com
Emerson Process Management
EPM Building
P. O. Box 17033
Jebe Ali Free Zone
Dubai, United Arab Emirates
T 971.4.8835235
F 971.4.8835312
WARRANTY
Seller warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufactured
or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expiration of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen
(18) months from the date of shipment by Seller, whichever period expires first. Consumables, such as glass electrodes,
membranes, liquid junctions, electrolyte, o-rings, catalytic beads, etc., and Services are warranted for a period of 90
days from the date of shipment or provision.
Products purchased by Seller from a third party for resale to Buyer ("Resale Products") shall carry only the warranty extended by
the original manufacturer. Buyer agrees that Seller has no liability for Resale Products beyond making a reasonable commercial
effort to arrange for procurement and shipping of the Resale Products.
If Buyer discovers any warranty defects and notifies Seller thereof in writing during the applicable warranty period, Seller shall, at
its option, promptly correct any errors that are found by Seller in the firmware or Services, or repair or replace F.O.B. point of manufacture that portion of the Goods or firmware found by Seller to be defective, or refund the purchase price of the defective portion of the Goods/Services.
All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuitable environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any other
cause not the fault of Seller are not covered by this limited warranty, and shall be at Buyer's expense. Seller shall not be obligated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by
an authorized Seller representative. All costs of dismantling, reinstallation and freight and the time and expenses of Seller's personnel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Seller.
Goods repaired and parts replaced during the warranty period shall be in warranty for the remainder of the original warranty period or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Seller and can be amended only
in a writing signed by an authorized representative of Seller. Except as otherwise expressly provided in the Agreement, THERE
ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESS OR IMPLIED, AS TO MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES.
RETURN OF MATERIAL
Material returned for repair, whether in or out of warranty, should be shipped prepaid to:
Emerson Process Management
2400 Barranca Parkway
Irvine, CA 92606
The shipping container should be marked:
Return for Repair
Model _______________________________
The returned material should be accompanied by a letter of transmittal which should include the following information (make a
copy of the "Return of Materials Request" found on the last page of the Manual and provide the following thereon):
1.
2.
3.
4.
5.
Location type of service, and length of time of service of the device.
Description of the faulty operation of the device and the circumstances of the failure.
Name and telephone number of the person to contact if there are questions about the returned material.
Statement as to whether warranty or non-warranty service is requested.
Complete shipping instructions for return of the material.
Adherence to these procedures will expedite handling of the returned material and will prevent unnecessary additional charges
for inspection and testing to determine the problem with the device.
If the material is returned for out-of-warranty repairs, a purchase order for repairs should be enclosed.
The right people,
the right answers,
right now.
ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
http://www.rosemountanalytical.com
Specifications subject to change without notice.
8
Credit Cards for U.S. Purchases Only.
Emerson Process Management
2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 757-8500
Fax: (949) 474-7250
http://www.rosemoutanalytical.com
© Rosemount Analytical Inc. 2012