Emerson 1057 THREE-INPUT INTELLIGENT ANALYZER Instruction Manual
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Instruction Manual
PN 51-1057/rev.C
January 2010
THREE-INPUT INTELLIGENT ANALYZER
Model 1057
ESSENTIAL INSTRUCTIONS
WARNING
READ THIS PAGE BEFORE PROCEEDING!
Your instrument purchase from Rosemount
Analytical, Inc. is one of the finest available for your
particular application. These instruments have been
designed, and tested to meet many national and
international standards. Experience indicates that its
performance is directly related to the quality of the
installation and knowledge of the user in operating
and maintaining the instrument. To ensure their continued operation to the design specifications, personnel should read this manual thoroughly before
proceeding with installation, commissioning, operation, and maintenance of this instrument. If this
equipment is used in a manner not specified by the
manufacturer, the protection provided by it against
hazards may be impaired.
• Failure to follow the proper instructions may
cause any one of the following situations to
occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty
invalidation.
• Ensure that you have received the correct model
and options from your purchase order. Verify that
this manual covers your model and options. If
not, call 1-800-854-8257 or 949-757-8500 to
request correct manual.
• For clarification of instructions, contact your
Rosemount representative.
• Follow all warnings, cautions, and instructions
marked on and supplied with the product.
• Use only qualified personnel to install, operate,
update, program and maintain the product.
• Educate your personnel in the proper installation,
operation, and maintenance of the product.
• Install equipment as specified in the Installation
section of this manual. Follow appropriate local
and national codes. Only connect the product to
electrical and pressure sources specified in this
manual.
• Use only factory documented components for
repair. Tampering or unauthorized substitution of
parts and procedures can affect the performance
and cause unsafe operation of your process.
• All equipment doors must be closed and protective covers must be in place unless qualified personnel are performing maintenance.
Emerson Process Management
Liquid Division
2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 757-8500
Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2009
RISK OF ELECTRICAL SHOCK
Equipment protected throughout by double insulation.
• Installation and servicing of this product may expose personel
to dangerous voltages.
• Main power wired to separate power source must be
disconnected before servicing.
• Do not operate or energize instrument with case open!
• Signal wiring connected in this box must be rated at least
240 V.
• Non-metallic cable strain reliefs do not provide grounding
between conduit connections! Use grounding type bushings
and jumper wires.
• Unused cable conduit entries must be securely sealed by
non-flammable closures to provide enclosure integrity in
compliance with personal safety and environmental protection
requirements. Unused conduit openings must be sealed with
NEMA 4X or IP65 conduit plugs to maintain the ingress
protection rating (NEMA 4X).
• Electrical installation must be in accordance with the National
Electrical Code (ANSI/NFPA-70) and/or any other applicable
national or local codes.
• Operate only with front panel fastened and in place.
• Proper use and configuration is the responsibility of the
user.
CAUTION
This product generates, uses, and can radiate radio frequency
energy and thus can cause radio communication interference.
Improper installation, or operation, may increase such interference. As temporarily permitted by regulation, this unit has not
been tested for compliance within the limits of Class A computing devices, pursuant to Subpart J of Part 15, of FCC Rules,
which are designed to provide reasonable protection against
such interference. Operation of this equipment in a residential
area may cause interference, in which case the user at his own
expense, will be required to take whatever measures may be
required to correct the interference.
CAUTION
This product is not intended for use in the light industrial,
residential or commercial environments per the instrument’s certification to EN50081-2 (EN61000-6-2).
QUICK START GUIDE
Model 1057 Three Input Analyzer
1. Refer to Section 2.0 for mechanical installation instructions.
2.
Wire sensor(s) to the signal boards. See Section 3.0 for wiring instructions. Refer to the sensor instruction
sheet for additional details. Make current output, alarm relay and power connections.
3. Once connections are secured and verified, close panel and apply power to the analyzer.
WARNING
WARNING
RISK OF ELECTRICAL SHOCK
Electrical installation must be in accordance with the
National Electrical Code (ANSI/NFPA-70) and/or any
other applicable national or local codes.
4. When the analyzer is powered up for the first time, Quick Start screens appear. Quick Start operating tips
are as follows:
a. A backlit field shows the position of the cursor.
b. To move the cursor left or right, use the keys to the left or right of the ENTER key. To scroll up or down
or to increase or decrease the value of a digit use the keys above and below the ENTER key . Use the
left or right keys to move the decimal point.
c. Press ENTER to store a setting. Press EXIT to leave without storing changes. Pressing EXIT during Quick
Start returns the display to the initial start-up screen (select language).
5. Complete the steps as shown in the Quick Start Guide flow diagram, Fig. A on the following page.
6. After the last step, the main display appears. The outputs are assigned to default values.
7.
To change output, and temperature-related settings, go to the main menu and choose Program. Follow the
prompts. For a general guide to the Program menu, see the Quick Reference Guide, Fig.B.
8. To return the analyzer to the default settings, choose Reset Analyzer under the Program menu.
D
Figure A. QUICK START GUIDE
QUICK START GUIDE – MODEL 1057
E
Figure B. MODEL 1057 MENU TREE
QUICK REFERENCE GUIDE
About This Document
This manual contains instructions for installation and operation of the Model 1057 Three-Input Intelligent Analyzer.
The following list provides notes concerning all revisions of this document.
F
Rev. Level
A
Date
01/09
B
C
10/09
01/10
Notes
This is the initial release of the product manual. The manual has been reformatted
to reflect the Emerson documentation style and updated to reflect any changes
in the product offering.
Russian language added. UL Approval added.
CSA Class I, Div 2 Non-incendive harardous area approval added.
MODEL 1057
TABLE OF CONTENTS
MODEL 1057
THREE INPUT INTELLIGENT ANALYZER
TABLE OF CONTENTS
QUICK START GUIDE
QUICK REFERENCE GUIDE
TABLE OF CONTENTS
Section Title
1.0
DESCRIPTION AND SPECIFICATIONS ................................................................
2.0
INSTALLATION .......................................................................................................
2.1
Unpacking and Inspection........................................................................................
2.2
Installation................................................................................................................
3.0
WIRING....................................................................................................................
3.1
General ....................................................................................................................
3.2
Preparing Conduit Openings....................................................................................
3.3
Preparing Sensor Cable ..........................................................................................
3.4
Power, Output, Alarms and Sensor Connections.....................................................
4.0
DISPLAY AND OPERATION ...................................................................................
4.1
User Interface ..........................................................................................................
4.2
Instrument Keypad...................................................................................................
4.3
Main Display ............................................................................................................
4.4
Menu System ...........................................................................................................
5.0
PROGRAMMING THE ANALYZER – BASICS .......................................................
5.1
5.2
5.3
5.4
5.5
5.6
5.7
5.8
5.9
6.0
6.1
6.2
6.3
6.4
7.0
7.1
7.2
7.3
7.4
7.5
8.0
General ....................................................................................................................
Changing StartUp Settings ......................................................................................
Choosing Temperature Units and Automatic/Manual Temperature Compensation .
Configuring and Ranging the Current Outputs.........................................................
Setting a Security Code ...........................................................................................
Security Access........................................................................................................
Using Hold ...............................................................................................................
Resetting Factory Default Settings ........................................................................
Programming Alarm Relays .....................................................................................
PROGRAMMING - MEASUREMENTS ...................................................................
Programming Measurements – Introduction ...........................................................
pH ............................................................................................................................
ORP .........................................................................................................................
Contacting Conductivity ..........................................................................................
CALIBRATION ......................................................................................................
Calibration – Introduction .........................................................................................
pH Calibration ..........................................................................................................
ORP Calibration .......................................................................................................
Contacting Conductivity Calibration .........................................................................
Calibrating Temperature ..........................................................................................
RETURN OF MATERIAL ........................................................................................
Warranty...................................................................................................................
Ordering Information ................................................................................................
i
Page
1
5
5
5
9
9
10
10
10
15
15
15
16
17
19
19
19
20
20
22
23
23
24
25
29
29
30
31
33
39
39
39
42
43
46
51
51
53
MODEL 1057
TABLE OF CONTENTS
TABLE OF CONTENTS CONT’D
LIST OF FIGURES
Fig#
A
B
2-1
2-2
3-1
3-2
3-3
3-4
3-5
3-6
3-7
3-8
3-9
4-1
5-1
5-2
5-3
5-4
5-5
6-1
6-2
7-1
7-2
7-3
7-4
Section
PREFACE
PREFACE
SEC 2.0
SEC 2.0
SEC 3.1
SEC 3.4
SEC 3.4
SEC 3.4
SEC 3.4
SEC 3.4
SEC 3.4
SEC 3.4
SEC 3.4
SEC 4.3
SEC 5.3.2
SEC 5.4.5
SEC 5.5.2
SEC 5.7.2
SEC 5.8.2
SEC 6.4
SEC 6.4
SEC 7.2
SEC 7.3
SEC 7.4
SEC 7.9
Figure Title
Quick Start Guide
Quick Reference Guide
Panel Mounting Dimensions ..............................................................
Pipe and Wall Mounting Dimensions .................................................
Connectors and Signal Input boards .................................................
24VDC Power Supply ........................................................................
Switching AC Power Supply...............................................................
Alarm Relay Wiring ............................................................................
Contacting Conductivity board and sensor cable leads .....................
pH/ORP/ISE signal board and sensor cable leads ............................
Power Wiring for Model 1057 85-265 VAC ........................................
Output Wiring for Model 1057 Main PCB...........................................
Power Wiring for Model 1057 24VDC ................................................
Formatting the Main Display .............................................................
Choosing Temp Units and Manual Auto Temp Compensation ...........
Configuring and Ranging the Current Outputs...................................
Setting A Security Code ....................................................................
Using Hold .........................................................................................
Resetting Factory Default Settings ....................................................
Configuring pH/ORP Measurements .................................................
Configure Contacting Measurements ...............................................
Calibrate pH .......................................................................................
Calibrate ORP....................................................................................
Calibrate Contacting Conductivity......................................................
Calibrate Temperature .......................................................................
Page
6
7
9
10
10
11
12
12
13
13
14
18
20
21
22
23
24
36
37
47
48
49
50
LIST OF TABLES
Number
5-1
6-1
6-2
6-3
7-1
7-2
7-3
7-11
Section
SEC 5.2.1
SEC 6.2.1
SEC 6.3.1
SEC 6.4.1
SEC 7.2
SEC 7.3
SEC 7.4
SEC 7.5
Table Title
Measurements and Measurement Units ...........................................
pH Measurement Programming .........................................................
ORP Measurement Programming ......................................................
Contacting Conductivity Measurement Programming........................
pH Calibration Routines .......................................................................
ORP Calibration Routine...................................................................
Contacting Conductivity Calibration Routines......................................
Temperature Calibration Routines........................................................
ii
Page
19
30
31
33
39
42
43
46
MODEL 1057
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
SECTION 1.0.
DESCRIPTION AND SPECIFICATIONS
• MULTI-PARAMETER INSTRUMENT – up to three inputs. Choose pH/ORP/ISE or Contacting
Conductivity/Resistivity in any combination.
• LARGE DISPLAY – large easy-to-read process measurements.
• EASY TO INSTALL – modular boards, removable connectors, easy to wire power, sensors, and outputs.
• INTUITIVE MENU SCREENS with advanced diagnostics and help screens.
• SMART -Enabled - compatible with SMART pH sensors.
• EIGHT Menu Languages.
• FOUR ANALOG OUTPUTS.
• UL and CSA APPROVED.
FEATURES AND APPLICATIONS
The Model 1057 analyzer offers three sensor inputs and
four current outputs thus reducing the cost per loop and
saving panel space. The pH signal input board
supports pH, ORP, and Ion-Selective Electrode measurements. The Conductivity signal input board supports
contacting conductivity, resistivity, Total Dissolved
Solids, salinity and percent concentration curves for
special applications. The modular design allows signal
input boards to be field replaced making configuration
changes easy. Conveniently, live process values are
always displayed during programming and calibration
routines. Standard features include isolated inputs,
8embedded local languages, four 4-20mA current
outputs, 4 alarm relays and removable connectors
for power and current outputs.
QUICK START PROGRAMMING: Exclusive Quick
Start screens appear the first time the Model 1057
is powered. The instrument auto-recognizes each
measurement board and prompts the user to configure
each sensor loop in a few quick steps for immediate
deployment.
MENUS: Menu screens for calibrating and programming
are simple and intuitive. Plain language prompts and help
screens guide the user through these procedures.
4-ELECTRODE CONDUCTIVITY: For applications
requiring wide range conductivity measurements, use
Rosemount Analytical’s Model 410VP PUR-SENSE
4-electrode sensor. It is not affected by fouling and is
supported by the same contacting conductivity signal
board as traditional 2-electrode sensors.
THREE SENSOR INPUTS: The Model 1057 accepts
one, two or three isolated inputs. Inputs are isolated from other signal sources and earth ground.
FOUR CURRENT OUTPUTS: Four 0/4-20 mA current outputs are electrically isolated. Outputs are
fully scalable and can be programmed to linear or
logrithmic modes. Output dampening can be enabled
with time constants from 0 to 999 seconds.
ENCLOSURE: The instrument fits standard ½ DIN
panel cutouts. The versatile enclosure design supports
panel-mount, pipe-mount, and surface/wall-mount
installations.
SMART pH SENSORS: Avoid buffer calibrations
in the field. Use Rosemount Analytical SMART pH
sensors to automatically calibrate the measurement loop when connected to Model 1057. Choose
from a complete range of SMART pH sensors suited to most applications.
1
MODEL 1057
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
SECURITY ACCESS CODES: Two levels of security
access are available. Program one access code for
routine calibration and hold of current outputs; program
another access code for all menus and functions.
Approvals:
RFI/EMI: EN-61326
LVD:
EN-61010-1
DIAGNOSTICS: The analyzer continuously monitors
itself and the sensor(s) for problematic conditions.
The display flashes Fault and/or Warning when these
conditions occur.
Hazardous Location Approvals:
Ordering options for CSA: all except UL
Diagnostics
Faults
Warnings
Sensor 1
Sensor 2
Sensor 3
Out 1: 12.05 mA
Out 2: 12.05 mA
Out 3: 12.05 mA
Out 4: 12.05 mA
1057-02-20-30-42
Instr SW VER: 3.12
AC Freq. Used: 60Hz
Information about
each condition
is quickly accessible
by pressing DIAG on
the keypad. User
help screens are
displayed for most
fault and warning
conditions to assist in
troubleshooting.
DISPLAY: The high-contrast LCD provides live
measurement readouts in large digits and shows up to
six additional process variables or diagnostic parameters. The display is back-lit and the format can be customized to meet user requirements.
LOCAL LANGUAGES : Rosemount Analytical extends
its worldwide reach by offering eight local languages –
English, French, German, Italian, Spanish,
Portuguese, Chinese and Russian. Every unit includes
user programming menus; calibration routines; faults
and warnings; and user help screens in all eight languages. The displayed language can be easily set and
changed using the menus.
SPECIFICATIONS - General
Enclosure: Polycarbonate. NEMA 4X/CSA 4 (IP65).
Dimensions: Overall 155 x 155 x 131mm (6.10 x 6.10 x
5.15 in.). Cutout: 1/2 DIN 139mm x 139mm (5.45 x 5.45 in.)
Minimum depth for panel mount insatllations 101.6 mm
(4.0 in).
Conduit Openings: Accepts 1/2” or PG13.5 conduit
fittings
Display: Monochromatic graphic liquid crystal display.
128 x 96 pixel display resolution. Backlit. Active
display area: 58 x 78mm (2.3 x 3.0 in.).
Ambient Temperature and Humidity: 0 to 55°C
(32 to 131°F). RH 5 to 95% (non-condensing)
Storage Temperature Effect: -20 to 60ºC (-4 to 140°F)
Power: Code -02: 20 to 30 VDC. 15 W.
Code -03: 84 to 265 VAC, 47.5 to 65.0 Hz, switching.
15 W.
Equipment protected by double insulation
2
Class I, Division 2, Groups A, B, C, & D
Class Il, Division 2, Groups E, F, & G
Class Ill T4A Tamb= 50° C
Evaluated to the ANSI/UL Standards. The ‘C’ and ‘US’ indicators adjacent to the CSA Mark signify that the product
has been evaluated to the applicable CSA and ANSI/UL
Standards, for use in Canada and the U.S. respectively.
Ordinary Locations: (only with -UL ordering option)
POLLUTION DEGREE 2: Normally only non-conductive
pollution occurs. Occasionally, however, a temporary
conductivity caused by condensation must be expected.
Altitude: for use up to 2000 meters (6562 ft.)
Alarms relays: Four alarm relays for process measurement(s) or temperature. Any relay can be configured as a fault alarm instead of a process alarm. Each
relay can be configured independently and each can
be programmed with interval timer settings.
Maximum Relay Current
Resistive
28 VDC
115 VAC
230 VAC
5.0 A
5.0 A
5.0 A
CAUTION
RISK OF ELECTRICAL SHOCK
Relays: Form C, SPDT, epoxy sealed
Inductive load: 1/8 HP motor (max.), 40 VAC
Inputs: Up to three sensor inputs-electrically isolated
Outputs: Four 4-20 mA or 0-20 mA isolated current
outputs. Fully scalable. Max Load: 550 Ohms.
Current Output Accuracy: ±0.05 mA @25 ºC
Terminal Connections Rating: Power connector
(3-leads): 24-12 AWG wire size. Signal board terminal blocks: 26-16 AWG wire size. Current output
connectors (4-leads): 24-16 AWG wire size. Alarm
relay terminal blocks: 24-12 AWG wire size
Weight/Shipping Weight: (rounded up to nearest lb or
nearest 0.5 kg): 3 lbs/4 lbs (1.5 kg/2.0 kg)
MODEL 1057
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
CONTACTING CONDUCTIVITY (Codes -20, -30 and -40)
Measures conductivity in the range 0 to 600,000 μS/cm
(600mS/cm). Measurement choices are conductivity,
resistivity, total dissolved solids, salinity, and % concentration. The % concentration selection includes the
choice of five common solutions (0-12% NaOH, 0-15%
HCl, 0-20% NaCl, and 0-25% or 96-99.7% H2SO4).
Temperature Specifications:
The conductivity concentration algorithms for these
solutions are fully temperature compensated. Three
temperature compensation options are available:
manual slope (X%/°C), high purity water (dilute sodium
chloride), and cation conductivity (dilute hydrochloric
acid). Temperature compensation can be disabled,
allowing the analyzer to display raw conductivity. For
more information concerning the use and operation of
the contacting conductivity sensors, refer to the product
data sheets.
Note: When contacting conductivity sensors are used
for sensor 1 and sensor 2, Model 1057 can derive an
inferred pH value called pHCalc. pHCalc is calculated
pH, not directly measured pH.
Note: Selected 4-electrode, high-range contacting
conductivity sensors are compatible with Model 1056.
Input filter: time constant 1 - 999 sec, default 2 sec.
Response time: 3 seconds to 100% of final reading
Salinity: uses Practical Salinity Scale
Temperature range
0-200ºC
Temperature Accuracy,
Pt-1000, 0-50 ºC
± 0.1ºC
Temperature Accuracy,
Pt-1000, Temp. > 50 ºC
± 0.5ºC
RECOMMENDED SENSORS FOR CONDUCTIVITY:
All Rosemount Analytical ENDURANCE Model 400
series conductivity sensors (Pt 1000 RTD) and
Model 410 sensor.
family
4-electrode sensors
Total Dissolved Solids: Calculated by multiplying
conductivity at 25ºC by 0.65
ENDURANCE series of
conductivity sensors
TM
PERFORMANCE SPECIFICATIONS
Recommended Range – Contacting Conductivity
Cell
Constant
0.01
0.1
1.0
4-electrode
0.01μS/cm
0.1μS/cm
1.0μS/cm
10μS/cm
100μS/cm
0.01μS/cm to 200μS/cm
1000μS/cm
10mS/cm
100mS/cm
1000mS/cm
200μS/cm to 6000μS/cm
2000μS/cm to 60mS/cm
0.1μS/cm to 2000μS/cm
1 μS/cm to 20mS/cm
20mS/cm to 600mS/cm
2 μS/cm to 300mS/cm
Cell Constant Linearity
±0.6% of reading in recommended range
+2 to -10% of reading outside high recommended range
±5% of reading outside low recommended range
±4% of reading in recommended range
3
MODEL 1057
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
pH/ORP/ISE (Codes -22, -32 and -42)
For use with any standard pH or ORP sensor.
Measurement choices are pH, ORP, Redox, ammonia,
fluoride or custom ISE. The automatic buffer recognition
feature uses stored buffer values and their temperature
curves for the most common buffer standards available
worldwide. The analyzer will recognize the value of the
buffer being measured and perform a self stabilization
check on the sensor before completing the calibration.
Manual or automatic temperature compensation is
menu selectable. Change in pH due to process temperature can be compensated using a programmable temperature coefficient. For more information concerning
the use and operation of the pH or ORP sensors, refer
to the product data sheets.
PERFORMANCE SPECIFICATIONS ANALYZER (ORP INPUT)
Measurement Range [ORP]: -1500 to +1500 mV
Accuracy: ± 1 mV
Temperature coefficient: ±0.12mV / ºC
Input filter: Time constant 1 - 999 seconds, default 4
seconds.
Response time: 5 seconds to 100% of final reading
RECOMMENDED SENSORS FOR ORP:
Model 1057 can also derive an inferred pH value called All standard ORP sensors.
pHCalc (calculated pH). pHCalc can be derived and
displayed when two contacting conductivity sensors are Rosemount Analytical pH Sensors
used as sensor 1 and sensor 2.
PERFORMANCE SPECIFICATIONS ANALYZER (pH INPUT)
Measurement Range [pH]: 0 to 14 pH
Accuracy: ±0.01 pH
Diagnostics: Glass impedance, reference impedance
Temperature coefficient: ±0.002pH/ ºC
Solution temperature correction: Pure water, dilute
base and custom.
Buffer recognition: NIST, DIN 19266, JIS 8802, and BSI.
Input filter: Time constant 1 - 999 seconds, default 4
seconds.
Response time: 5 seconds to 100%
Sensor Compatibility: Model 1057 is also compatible
with SMART pH sensors from Rosemount
Analytical. Choose from 14 SMART sensor models
for a wide array of applications and process conditions.
Temperature Specifications:
Temperature range
0-150ºC
Temperature Accuracy, Pt-100, 0-50 ºC
± 0.5ºC
Temperature Accuracy, Temp. > 50 ºC
± 1ºC
4
Fourteen SMART sensor models include PERpH-X, TUpH and
general purpose pH sensor families.
MODEL 1057
SECTION 2.0
INSTALLATION
SECTION 2.0.
INSTALLATION
2.1 UNPACKING AND INSPECTION
2.2 INSTALLATION
2.1 UNPACKING AND INSPECTION
Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions. Save the box. If
there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present. If
items are missing, notify Rosemount Analytical immediately.
2.2 INSTALLATION
2.2.1 General Information
1. Although the analyzer is suitable for outdoor use, do not install it in direct sunlight or in areas of extreme temperatures.
2. Install the analyzer in an area where vibration and electromagnetic and radio frequency interference are minimized or absent.
3. Keep the analyzer and sensor wiring at least one foot from high voltage conductors. Be sure there is easy
access to the analyzer.
4. The analyzer is suitable for panel, pipe, or surface mounting. Refer to the table below.
Type of Mounting
Figure
Panel
2-1
Wall and Pipe
2-2
WARNING
RISK OF ELECTRICAL SHOCK
Electrical installation must be in accordance with the National Electrical Code
(ANSI/NFPA-70) and/or any other applicable national or local codes.
5
FIGURE 2-1 PANEL MOUNTING DIMENSIONS
MILLIMETER
INCH
Front View
Side View
Bottom View
The front panel is hinged at the bottom. The panel swings down for easy access to the wiring locations.
Panel mounting seal integrity (4/4X) for outdoor applications is the responsibility of the end user.
6
FIGURE 2-2 PIPE AND WALL MOUNTING DIMENSIONS
(Mounting bracket PN:23820-00)
MILLIMETER
INCH
154.9
6.1
Wall / Surface Mount
232
9.1
102
4.0
33.5
1.3
130
5.1
187
7.4
154.9
6.1
165
6.5
Side View
Front View
Pipe Mount
232
9.1
Bottom View
33.5
1.3
130
5.1
80.01
3.2
165
6.5
45.21
1.8
108.9
4.3
Side View
71.37
2.8
The front panel is hinged at the bottom. The panel swings down for easy access to the wiring locations.
7
MODEL 1057
8
SECTION 2.0
INSTALLATION
This page left blank intentionally
MODEL 1057
SECTION 3.0
WIRING
SECTION 3.0.
WIRING
3.1
3.2
3.3
3.4
GENERAL
PREPARING CONDUIT OPENINGS
PREPARING SENSOR CABLE
POWER, OUTPUT, AND SENSOR
CONNECTIONS
3.1 GENERAL
The Model 1057 is easy to wire. It includes removable connectors and slide-out signal input boards.
3.1.1 Removable connectors and signal input boards
Model 1057 uses removable signal input boards and communication boards for ease of wiring and installation. Each of the signal input boards can be partially or completely removed from the enclosure for wiring.
The Model 1057 has three slots for placement of up to three signal input boards and one communication
board.
Slot 1 – Left
Signal board 3
Slot 2 – Center
Signal board 2
Slot 3 – Right
Signal board 1
Figure 3-1
Note: If the -UL option code has been ordered, a plastic insulator shield surrounds the entire power supply board
(AC power supply only). The protective insulator shield does not appear in this photo.
3.1.2 Signal Input boards
Slots 1, 2 and 3 are for signal input measurement boards. Wire the sensor leads to the measurement board
following the lead locations marked on the board. After wiring the sensor leads to the signal board, carefully slide
the wired board fully into the enclosure slot and take up the excess sensor cable through the cable gland. Tighten
the cable gland nut to secure the cable and ensure a sealed enclosure.
Note: that signal input board 3 is inserted into slot 1. Board 3 is inverted in the slot to allow board components to
face to the right. Board 3 uses a long ribbon cable to connect to the main PCB. Boards 1 and 2 use a split ribbon
cable to connect both signal boards to a common connector on the main board.
3.1.3 Alarm relays
Four alarm relays are supplied with the switching power supply (84 to 265VAC, -03 order code) and the 24VDC
power supply (20-30VDC, -02 order code). All relays can be used for process measurement(s) or temperature.
Any relay can be configured as a fault alarm instead of a process alarm. Each relay can be configured
independently and each can be programmed as an interval timer, typically used to activate pumps or control
valves. As process alarms, alarm logic (high or low activation or USP*) and deadband are user-programmable.
Customer-defined failsafe operation is supported as a programmable menu function to allow all relays to be
energized or not-energized as a default condition upon powering the analyzer.
The USP alarm can be programmed to activate when the conductivity is within a user-selectable
percentage of the limit. USP alarming is available only when a contacting conductivity measurement board is
installed.
9
MODEL 1057
SECTION 3.0
WIRING
3.2 PREPARING CONDUIT OPENINGS
There are six conduit openings in all configurations of Model 1057. (Note that four plugs are provided upon shipment.)
Conduit openings accept 1/2-inch conduit fittings or PG13.5 cable glands. To keep the case watertight, block
unused openings with NEMA 4X or IP65 conduit plugs.
NOTE: Use watertight fittings and hubs that comply with your requirements. Connect the conduit hub to the
conduit before attaching the fitting to the analyzer.
3.3 PREPARING SENSOR CABLE
The Model 1057 is intended for use with all Rosemount Analytical pH/ORP and contacting conductivity sensors. Refer
to the sensor installation instructions for details on preparing sensor cables.
3.4 POWER, OUTPUT, AND SENSOR CONNECTIONS
All field wiring must be rated for 75ºC or higher. Each instrument includes a printed label inside the enclosure stating
this wiring requirement.
3.4.1 Power wiring
Two Power Supplies are offered for Model 1057:
a. 24VDC (20 – 30V) Power Supply (-02 ordering code)
b. 84 – 265 VAC Switching Power Supply (-03 ordering code)
AC mains (115 or 230V) leads and 24VDC leads are wired to the Power Supply board which is mounted vertically
on the left side of the main enclosure cavity. Each lead location is clearly marked on the Power Supply board.
Wire the power leads to the Power Supply board using the lead markings on the board.
24VDC Power Supply (-02 ordering code) is shown below:
This power supply automatically
detects DC power and accepts
20VDC to 30VDC inputs.
Four programmable alarm relays are
included.
Figure 3-2
Switching AC Power Supply (-03 ordering code) is shown below:
This power supply automatically
detects AC line conditions and switches
to the proper line voltage and line
frequency.
Four programmable alarm relays are
included.
Figure 3-3
10
MODEL 1057
SECTION 3.0
WIRING
3.4.2 Current Output wiring
All instruments are shipped with four 4-20mA current outputs. Wiring locations for the outputs are on the Main
board which is mounted on the hinged door of the instrument. Wire the output leads to the correct position on the
Main board connectors using the lead markings (+/positive, -/negative) on the board. Male mating connectors are provided with each unit. Use a 3/32”-wide standard blade screwdriver.
3.4.3 Alarm relay wiring
Four alarm relays are supplied with the switching power supply (84 to 265VAC, -03 order code) and the 24VDC
power supply (20-30VDC, -02 order code). Wire the relay leads on each of the independent relays to the correct
position on the power supply board using the printed lead markings (NO/Normally Open, NC/Normally Closed, or
Com/Common) on the board.
NO1
COM1
RELAY 1
NC1
NO2
COM2
RELAY 2
NC2
NO3
COM3
RELAY 3
NC3
NO4
COM4
RELAY 4
NC4
Figure 3-4 Alarm Relay Wiring for Model 1057 Switching Power Supply (-03 Order Code)
WARNING
RISK OF ELECTRICAL SHOCK
Electrical installation must be in accordance with
the National Electrical Code (ANSI/NFPA-70)
and/or any other applicable national or local codes.
3.4.4 Sensor wiring to signal boards
Wire the correct sensor leads to the measurement board using the lead locations marked directly on the b o a r d .
After wiring the sensor leads to the signal board, carefully slide the wired board fully into the enclosure slot and
take up the excess sensor cable through the cable gland.
For best EMI/RFI protection use shielded output signal cable enclosed in an earth-grounded metal conduit.
Connect the shield to earth ground. AC wiring should be 14 gauge or greater. Provide a switch or breaker to disconnect the analyzer from the main power supply. Install the switch or breaker near the analyzer and label it as
the disconnecting device for the analyzer.
Keep sensor and output signal wiring separate from power wiring. Do not run sensor and power wiring in the same
conduit or close together in a cable tray.
11
MODEL 1057
SECTION 3.0
WIRING
Sec. 3.4 Signal board wiring
Figure 3-5 Contacting Conductivity signal board and Sensor cable leads
Figure 3-6 pH/ORP/ISE signal board and Sensor cable leads
12
MODEL 1057
SECTION 3.0
WIRING
FIGURE 3-7 Power Wiring for Model 1057 84-265 VAC Power Supply (-03 ordering code)
FIGURE 3-8 Output Wiring for Model 1057 Main PCB
13
MODEL 1057
SECTION 3.0
WIRING
To Main PCB
FIGURE 3-9 Power Wiring for Model 1057 24VDC Power Supply (-02 ordering code)
14
MODEL 1057
SECTION 4.0
DISPLAY AND OPERATION
SECTION 4.0
DISPLAY AND OPERATION
4.1
4.2
4.3
4.4
USER INTERFACE
KEYPAD
MAIN DISPLAY
MENU SYSTEM
4.1 USER INTERFACE
The Model 1057 has a large display which shows
three live measurement readouts in large digits and up
to six additional process variables or diagnostic
parameters concurrently. The display is back-lit and the
format can be customized to meet user requirements.
The intuitive menu system allows access to Calibration,
Hold (of current outputs), Programming, and Display
functions by pressing the MENU button. In addition, a
dedicated DIAGNOSTIC button is available to provide
access to useful operational information on installed
sensor(s) and any problematic conditions that might
occur. The display flashes Fault and/or Warning when
these conditions occur. Help screens are displayed for
most fault and warning conditions to guide the user in
troubleshooting.
During calibration and programming, key presses cause
different displays to appear. The displays are selfexplanatory and guide the user step-by-step through
the procedure.
4.2 INSTRUMENT KEYPAD
There are 4 Function keys and 4 Selection keys on the
instrument keypad.
Function keys:
The MENU key is used to access menus for programming and calibrating the instrument. Four top-level
menu items appear when pressing the MENU key:
Calibrate: calibrate attached sensors and
analog outputs.
Hold: Suspend current outputs.
Program: Program outputs, measurement,
temperature, security and reset.
Display: Program display format, language,
warnings, and contrast
Pressing MENU always causes the main menu screen
to appear. Pressing MENU followed by EXIT causes
the main display to appear.
15
MODEL 1057
Pressing the DIAG key displays active Faults and
Warnings, and provides detailed instrument information
and sensor diagnostics including: Faults, Warnings,
Sensor 1, 2 and 3 information, Current Outputs live
values, model configuration string e.g. 1057-03 -20-3042-AN, Instrument Software version, and AC frequency. Pressing ENTER on Sensor 1 or Sensor 2 provides
useful diagnostics and information (as applicable):
Measurement, Sensor Type, Raw signal value, Cell
constant, Zero Offset and Temperature. Offset, selected
SECTION 4.0
DISPLAY AND OPERATION
measurement range, Cable Resistance, Temperature
Sensor Resistance, Signal Board software version.
The ENTER key. Pressing ENTER stores numbers and
settings and moves the display to the next screen.
The EXIT key. Pressing EXIT returns to the previous
screen without storing changes.
Selection keys:
Surrounding the ENTER key, four Selection keys – up,
down, right and left, move the cursor to all areas of the
screen while using the menus.
Selection keys are used to:
1. select items on the menu screens
2. scroll up and down the menu lists.
3. enter or edit numeric values.
4. move the cursor to the right or left
5. select measurement units during operations
4.3 MAIN DISPLAY
The Model 1057 displays one, two or three primary measurement values, up to six secondary measurement
values, a fault and warning banner, alarm relay flags.
Process measurements:
Three process variables are displayed if three signal
boards are installed. One process variable and
process temperature is displayed if one signal board is
installed with one sensor. The Upper display area
shows the Sensor 1 process reading. The Center display area shows the Sensor 2 process reading. For
dual conductivity, the display areas can be assigned to
different process variables as follows:
Process variables for display- examples:
Measure 1
Measure 2
Measure 3
% Reject
% Pass
Ratio
Blank
pH Calc
Secondary values:
Up to six secondary values are shown in six display
quadrants at the bottom of the screen. All four
secondary value positions can be programmed by the
user to any display parameter available. Possible
secondary values include:
Displayable Secondary Values
Slope 1, 2, 3
Output 1 mA
Ref Off 1, 2, 3
Output 2 mA
Gl Imp 1, 2, 3
Output 3 mA
Ref Imp 1, 2, 3
Output 4 mA
Raw 1, 2, 3
Output 1 %
mV Input 1, 2, 3
Output 2 %
Temp 1, 2, 3
Output 3 %
Man Temp 1, 2, 3
Output 4 %
Measure 1, 2, 3
Blank
16
MODEL 1057
SECTION 4.0
DISPLAY AND OPERATION
Fault and Warning banner:
If the analyzer detects a problem with itself or the sensor the word Fault or Warning will appear at the bottom of
the display. A fault requires immediate attention. A warning indicates a problematic condition or an impending failure. For troubleshooting assitance, press Diag.
Formatting the Main Display
The main display screen can be programmed to show primary process variables, secondary process variables and
diagnostics.
1.
Press MENU
2.
Scroll down to Display. Press ENTER.
3.
Main Format will be highlighted. Press ENTER.
4.
The sensor 1 process value will be highlighted in reverse video. Press the selection keys to navigate down
to the screen sections that you wish to program. Press ENTER.
5.
Choose the desired display parameter or diagnostic for each of the four display sections in the lower screen.
6.
Continue to navigate and program all desired screen sections.
return to the main display.
Press MENU and EXIT. The screen will
For single sensor configurations, the default display shows the live process measurement in the upper display area
and temperature in the center display area. The user can elect to disable the display of temperature in the center
display area using the Main Format function. See Fig. 4-1 to guide you through programming the main display to
select process parameters and diagnostics of your choice.
For dual sensor configurations, the default display shows Sensor 1 live process measurement in the display area
1 and Sensor 2 live process measurement temperature in the display area 2. See Fig. 4-1 to guide you through
programming the main display to select process parameters and diagnostics of your choice.
4.4 MENU SYSTEM
Model 1057 uses a scroll and select menu system.
Pressing the MENU key at any time opens the top-level
menu including Calibrate, Hold, Program and Display
functions.
To find a menu item, scroll with the up and down keys
until the item is highlighted. Continue to scroll and
select menu items until the desired function is chosen.
To select the item, press ENTER. To return to a previous menu level or to enable the main live display,
press the EXIT key repeatedly. To return immediately
to the main display from any menu level, simply press
MENU then EXIT.
The selection keys have the following functions:
The Up key (above ENTER) increments numerical values, moves the decimal place one place to the right,
or selects units of measurement.
The Down key (below ENTER) decrements numerical values, moves the decimal place one place to the
left, or selects units of measurement
The Left key (left of ENTER) moves the cursor to the left.
The Right key (right of ENTER) moves the cursor to the right.
To access desired menu functions, use the “Quick Reference” Figure B. During all menu displays (except main
display format and Quick Start), the live process measurements and secondary measurement values are
displayed in the top two lines of the Upper display area. This conveniently allows display of the live values during
important calibration and programming operations.
Menu screens will time out after two minutes and return to the main live display.
17
MODEL 1057
SECTION 4.0
DISPLAY AND OPERATION
FIGURE 4-1 Formatting the Main Display
18
MODEL 1057
SECTION 5.0
PROGRAMMING THE ANALYZER - BASICS
SECTION 5.0.
PROGRAMMING THE ANALYZER - BASICS
5.1
5.2
5.3
5.4
5.5
5.6
5.7
5.8
5.9
GENERAL
CHANGING START-UP SETTINGS
PROGRAMMING TEMPERATURE
CONFIGURING AND RANGING 4-20MA OUTPUTS
SETTING SECURITY CODES
SECURITY ACCESS
USING HOLD
RESETTING FACTORY DEFAULTS – RESET ANALYZER
PROGRAMMING ALARM RELAYS
5.1 GENERAL
Section 5.0 describes the following programming functions:
Changing the measurement type, measurement units and temperature units.
Choose temperature units and manual or automatic temperature compensation mode
Configure and assign values to the current outputs
Set a security code for two levels of security access
Accessing menu functions using a security code
Enabling and disabling Hold mode for current outputs
Choosing the frequency of the AC power (needed for optimum noise rejection)
Resetting all factory defaults, calibration data only, or current output settings only
5.2 CHANGING STARTUP SETTINGS
5.2.1 Purpose
To change the measurement type, measurement units, or temperature units that were initially entered in Quick
Start, choose the Reset analyzer function (Sec. 5.8) or access the Program menus for sensor 1, 2 or 3 (Sec. 6.0).
The following choices for specific measurement type, measurement units are available for each sensor measurement board.
TABLE 5-1. Measurements and Measurement Units
Signal board
Available measurements
Measurements units:
pH/ORP (-22, -32, -42)
pH, ORP, Redox, Ammonia, Fluoride,
Custom ISE
pH, mV (ORP)
%, ppm, mg/L, ppb, μg/L, (ISE)
Contacting conductivity
(-20, -30, -40)
Conductivity, Resistivity, TDS, Salinity,
NaOH (0-12%), HCl (0-15%), Low H2SO4,
High H2SO4, NaCl (0-20%), Custom Curve
μS/cm, mS/cm, S/cm
% (concentration)
Temperature (all)
Temperature
°C, ºF
5.2.2 Procedure.
Follow the Reset Analyzer procedure (Sec 5.8) to reconfigure the analyzer to display new measurements or
measurement units. To change the specific measurement or measurement units for each signal board type,
refer to the Program menu for the appropriate measurement (Sec. 6.0).
19
MODEL 1057
SECTION 5.0
PROGRAMMING THE ANALYZER - BASICS
5.3 CHOOSING TEMPERATURE UNITS AND AUTOMATIC/MANUAL TEMPERATURE
COMPENSATION
5.3.1 Purpose
Most liquid analytical measurements (except ORP)
require temperature compensation. The Model 1057
performs temperature compensation automatically by
applying internal temperature correction algorithms.
Temperature correction can also be turned off. If temperature correction is off, the Model 1057 uses the temperature entered by the user in all temperature correction calculations.
5.3.2 Procedure.
Follow the menu screens in Fig. 5.1 to select automatic
or manual temp compensation, set the manual
reference temperature, and to program temperature
units as °C or °F.
S1: 1.234µS/cm
S2: 12.34pH
S3:123.4µS/cm
123.4ºC
123.4ºC
123.4ºC
Temperature
Units:
°C
S1 Temp Comp: Auto
S2 Temp Comp: Auto
S3 Temp Comp: Auto
Figure 5-1. Choosing Temp Units and Manual Auto Temp Compensation
5.4 CONFIGURING AND RANGING THE CURRENT OUTPUTS
5.4.1 Purpose
The Model 1057 accepts inputs from three sensors and
has four analog current outputs. Ranging the outputs
means assigning values to the low (0 or 4 mA) and high
(20 mA) outputs. This section provides a guide for
configuring and ranging the outputs. ALWAYS
CONFIGURE THE OUTPUTS FIRST.
5.4.2 Definitions
1. CURRENT OUTPUTS. The analyzer provides a continuous output current (4-20 mA or 0-20 mA) directly
proportional to the process variable or temperature.
20
The low and high current outputs can be set to any
value.
2. ASSIGNING OUTPUTS. Assign a measurement to
Outputs 1, 2, 3, or 4.
3. DAMPEN. Output dampening smooths out noisy
readings. It also increases the response time of the
output. Output dampening does not affect the
response time of the display.
4. MODE. The current output can be made directly
proportional to the displayed value (linear mode) or
directly proportional to the common logarithm of the
displayed value (log mode).
MODEL 1057
SECTION 5.0
PROGRAMMING THE ANALYZER - BASICS
5.4.3 Procedure: Configure Outputs.
Under the Program/Outputs menu, the adjacent screen
will appear to allow configuration of the outputs. Follow
the menu screens in Fig. 5-2 to configure the outputs.
S1: 1.234µS/cm
S2: 12.34pH
S3:123.4µS/cm
123.4ºC
123.4ºC
123.4ºC
OutputM Configure
Assign:
S1 Meas
Range:
4-20mA
Scale:
Linear
Dampening:
0sec
Fault Mode: Fixed
Fault Value: 21.00mA
5.4.4 Procedure: Assigning Measurements the Low
and High Current Outputs
The adjacent screen will appear when entering the
Assign function under Program/Output/Configure.
These screens allow you to assign a measurement,
process value, or temperature input to each output.
Follow the menu screens in Fig. 5-2 to assign
measurements to the outputs.
5.4.5 Procedure: Ranging the Current Outputs
The
adjacent
screen
will
appear
under
Program/Output/Range. Enter a value for 4mA and
20mA (or 0mA and 20mA) for each output. Follow the
menu screens in Fig. 5-2 to assign values to the outputs.
S1: 1.234µS/cm
S2: 12.34pH
S3:123.4µS/cm
123.4ºC
123.4ºC
123.4ºC
OutputM Assign
S1 Measurement
S1 Temperature
S2 Measurement
S2 Temperature
S3 Measurement
S3 Temperature
S1: 1.234µS/cm
S2: 12.34pH
S3:123.4µS/cm
OM
OM
OM
OM
123.4ºC
123.4ºC
123.4ºC
Output Range
SN 4mA: 0.000µS/cm
SN 20mA: 20.00µS/cm
SN 4mA: 00.00pH
SN 20mA: 14.00pH
Figure 5-2. Configuring and Ranging the Current Outputs
21
MODEL 1057
SECTION 5.0
PROGRAMMING THE ANALYZER - BASICS
2.
3.
5.5 SETTING A SECURITY CODE
5.5.1 Purpose.
The security codes prevent accidental or unwanted
changes to program settings, displays, and calibration.
Model 1057 has two levels of security code to control
access and use of the instrument to different types of
users. The two levels of security are:
- All: This is the Supervisory security level. It
allows access to all menu functions, including
Programming, Calibration, Hold and Display.
- Calibration/Hold: This is the operator or technician level menu. It allows access to only
calibration and Hold of the current outputs.
4.
Scroll down to Security. Select Security.
The security entry screen appears. Enter a
three digit security code for each of the desired
security levels. The security code takes effect
two minutes after the last key stroke. Record
the security code(s) for future access and
communication to operators or technicians as
needed.
The display returns to the security menu
screen. Press EXIT to return to the previous
screen. To return to the main display, press
MENU followed by EXIT.
Fig. 5-3 displays the security code screens.
5.5.2 Procedure.
1. Press MENU. The main menu screen appears.
Choose Program.
Program
MAIN MENU
Figure 5-3. Setting a Security Code
S1: 1.234µS/cm 1.234ºC
S2: 12.34pH
1.234ºC
S3: 12.34µS/cm 1.234ºC
Program
Outputs
Measurement
Temperature
Security
Diagnostic Setup
Rejection Freq: 60Hz
Reset Analyzer
22
S1: 1.234µS/cm 1.234ºC
S2: 12.34pH
1.234ºC
S3: 12.34µS/cm 1.234ºC
Security
Calibration/Hold: 000
All: 000
MODEL 1057
SECTION 5.0
PROGRAMMING THE ANALYZER - BASICS
5.6 SECURITY ACCESS
5.7 USING HOLD
5.6.1 How the Security Code Works
When entering the correct access code for the
Calibration/Hold security level, the Calibration and
Hold menus are accessible. This allows operators or
technicians to perform routine maintenance. This
security level does not allow access to the Program or
Display menus.
When entering the correct access code for All security
level, the user has access to all menu functions, including Programming, Calibration, Hold and Display.
5.7.1 Purpose
The analyzer output is always proportional to measured
value. To prevent improper operation of systems or
pumps that are controlled directly by the current
output, place the analyzer in hold before removing
the sensor for calibration and maintenance. Be sure
to remove the analyzer from hold once calibration is
complete. During hold, both outputs remain at the last
value. Once in hold, all current outputs remain on
Hold indefinitely.
5.6.2 Procedure.
1. If a security code has been programmed, selecting
the Calibrate, Hold, Program or Display top menu
items causes the security access screen to appear
5.7.2 Using the Hold Function
To hold the outputs,
1.
Press MENU. The main menu screen appears.
Choose Hold.
2.
The Hold Outputs and Alarms? screen
appears. Choose Yes to place the analyzer in
hold. Choose No to take the analyzer out of
hold.
Note: There are no alarm relays with this con
figuration. Current outputs are included with all
configurations.
3.
The Hold screen will then appear and Hold
will remain on indefinitely until Hold is
disabled.
2. Enter the three-digit security code for the appropriate
security level.
S1: 1.234µS/cm 1.234ºC
S2: 12.34pH
1.234ºC
S3: 12.34µS/cm 1.234ºC
Security Code
000
3. If the entry is correct, the appropriate menu screen
appears. If the entry is incorrect, the Invalid Code
screen appears. The Enter Security Code screen
reappears after 2 seconds.
See figure 5-1 below.
Hold
MAIN MENU
Figure 5-4. Using Hold
S1: 1.234µS/cm 1.234ºC
S2: 12.34pH
1.234ºC
S3: 12.34µS/cm 1.234ºC
Hold
S1 Hold:
S2 Hold:
S3 Hold:
No
No
No
S1: 1.234µS/cm 1.234ºC
S2: 12.34pH
1.234ºC
S3: 12.34µS/cm 1.234ºC
S1 Hold outputs
and alarms?
No
Yes
23
MODEL 1057
SECTION 5.0
PROGRAMMING THE ANALYZER - BASICS
5.8 RESETTING FACTORY DEFAULT SETTINGS
5.8.1 Purpose.
This section describes how to restore factory calibration and default values. The process also clears all fault messages
and returns the display to the first Quick Start screen. The Model 1057 offers three options for resetting factory
defaults.
a. reset all settings to factory defaults
b. reset sensor calibration data only
c. reset output calibration only
5.8.2 Procedure.
To reset to factory defaults, reset calibration data only or reset analog outputs only, follow the Reset Analyzer flow
diagram.
Figure 5-5. Resetting Factory Default Settings
24
MODEL 1057
SECTION 5.0
PROGRAMMING THE ANALYZER - BASICS
5.9 Programming Alarm Relays
5.9.1 Purpose.
The Model 1057 24VDC (-02 order code) and the AC switching power supply (-03 order code) provide four alarm
relays for process measurement or temperature. Each alarm can be configured as a fault alarm instead of a
process alarm. Also, each relay can be programmed independently and each can be programmed as an interval
timer. This section describes how to configure alarm relays, simulate relay activation, and synchronize timers for
the four alarm relays. This section provides details to program the following alarm features:
Sec.
Alarm relay feature:
default
Description
5.9.2
Enter Setpoint
100.0uS/cm
Enter alarm trigger value
5.9.3
Assign measurement S1 Measure
Select alarm assignment
5.9.4
Set relay logic
Program relay to activate at High or Low reading
High
0.00uS/cm
Program the change in process value after the relay deactivates
5.9.5
Deadband:
5.9.6
USP Safety:
5.9.7
Normal state:
5.9.8
Interval time:
24.0 hr
Time in hours between relay activations
5.9.9
On-Time:
10 min
Enter the time in seconds that the relay is activated.
5.9.10
Recover time:
60 sec
Enter time after the relay deactivation for process recovery
5.9.11
Hold while active:
5.9.12
Simulate
5.9.13
Synchronize Timers
0%↓
Open
S1
Program percentage of the limit to activate the alarm
Program relay default condition as open or closed for failsafe operation
Holds current outputs during relay activation
Manually simulate alarms to confirm relay operation
Yes
Control the timing of two or more relay timers set as Interval timers
Under the Program/Alarms menu, this screen will
appear to allow configuration of the alarm relays.
Follow the menu screens in Fig. XX to configure the
outputs.
This screen will appear to allow selection of a specific
alarm relay. Select the desired alarm and press
ENTER.
This screen will appear next to allow complete programming of each alarm. Factory defaults are displayed as they would appear for an installed contacting conductivity board. USP Safety only appears if
alarm logic is set to “USP”. Interval timer, On Time,
Recover Time, and Hold While Active only appear if
the alarm is configured as an Interval timer.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
Alarms
Configure/Setpoint
Simulate
Synchronize Timers: Yes
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
Configure/Setpoint
Alarm 1
Alarm 2
Alarm 3
Alarm 4
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
AlarmM Settings
Setpoint:
100.0uS/cm
Assign:
S1 Measure
Logic:
High
Deadband: 0.00uS/cm
USP Safety:
0%↓
Interval time:
24.0 hr
On Time:
120 sec
Recover time:
60 sec
Hold while active: Sens1
25
MODEL 1057
5.9.2 Procedure – Enter Setpoints
Under the Program/Alarms menu, this screen will
appear to allow configuration of the alarm relays.
Enter the desired value for the process measurement
or temperature at which to activate an alarm event.
5.9.3 Procedure – Assign Measurement
Under the Alarms Settings menu, this screen will
appear to allow assignment of the alarm relays. select
an alarm assignment. Additional assignment choices
are shown in Figure X-X depending on which measurement board(s) is installed.
5.9.4 Procedure – Set Relay Logic
Under the Alarms Settings menu, this screen will
appear to set the alarm logic. Select the desired relay
logic to activate alarms at a High reading or a Low
reading. USP Safety only appears if a contacting conductivity board is installed.
5.9.5 Procedure – Deadband
Under the Alarms Settings menu, this screen will
appear to program the deadband as a measurement
value. Enter the change in the process value needed
after the relay deactivates to return to normal (and
thereby preventing repeated alarm activation).
SECTION 5.0
PROGRAMMING THE ANALYZER - BASICS
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
Alarm1 S2 Setpoint
+100.0uS/cm
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
AlarmM Assign:
S1 Measurement
S1 Temperature
S2 Measurement
S2 Temperature
S3 Measurement
S3 Temperature
Interval Timer
Fault
Off
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
AlarmM Logic:
High
Low
USP
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
Alarm1 Deadband
+000.5uS/cm
5.9.6 Procedure – USP Safety
Under the Alarms Settings menu, this screen will
appear to program the USP alarm setting. Enter the
percentage below the limit at which to activate the
alarm.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
5.9.7 Procedure – Normal state
The user can define failsafe condition in software by
programming the alarm default state to normally open
or normally closed upon power up. To display this
alarm configuration item, enter the Expert menus by
holding down the EXIT key for 6 seconds while in the
main display mode. Select Yes upon seeing the screen
prompt: “Enable Expert Menu?”
Under the Alarms Settings menu, this screen will
appear to set the normal state of the alarms. Select the
alarm condition that is desired each time the analyzer is
powering up.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
26
1.234ºC
1.234ºC
1.234ºC
1.234ºC
1.234ºC
1.234ºC
Alarm1 USP Safety
+0%
1.234ºC
1.234ºC
1.234ºC
Alarm2 Normal State
Open
Closed
MODEL 1057
5.9.8 Procedure – Interval time
Under the Alarms Settings menu, this screen will
appear to set the interval time. Enter the fixed time in
hours between relay activations.
5.9.9 Procedure – On time
Under the Alarms Settings menu, this screen will
appear to set the relay on time. Enter the time in seconds that the relay is activated.
5.9.10 Procedure – Recovery time
Under the Alarms Settings menu, this screen will
appear to set the relay recovery time. Enter time after
the relay deactivation for process recovery.
5.9.11 Procedure – Hold while active
Under the Alarms Settings menu, this screen will
appear to program the feature that Holds the current
outputs while alarms are active. Select to hold the
current outputs for Sensor 1, Sensor 2 or both sensors
while the relay is activated.
5.9.12 Procedure – Simulate
Alarm relays can be manually set for the purposes of
checking devices such as valves or pumps. Under the
Alarms Settings menu, this screen will appear to allow
manual forced activation of the alarm relays. Select
the desired alarm condition to simulate.
5.9.13 Procedure – Synchronize
Under the Alarms Settings menu, this screen will
appear to allow Synchronization of alarms that are set
to Interval Timers. Select yes or no to Synchronize
two or more timers.
SECTION 5.0
PROGRAMMING THE ANALYZER - BASICS
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
Alarm1 Interval Time
024.0 hrs
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
Alarm1 On-Time
00.00sec
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
Alarm1 Recovery
060sec
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
Alarm1 Hold while active
Sensor 1
Sensor 2
Sensor 3
All
None
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
Simulate Alarm M
Don’t simulate
De-energize
Energize
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
Synchronize Timers
Yes
No
27
MODEL 1057
28
SECTION 5.0
PROGRAMMING THE ANALYZER - BASICS
This page left blank intentionally
MODEL 1057
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
SECTION 6.0
PROGRAMMING - MEASUREMENTS
6.1
6.2
6.3
6.4
CONFIGURING MEASUREMENTS – INTRODUCTION
pH
ORP
CONTACTING CONDUCTIVITY
6.1 PROGRAMMING MEASUREMENTS – INTRODUCTION
The Model 1057 automatically recognizes each installed measurement board upon first power-up and each time
the analyzer is powered. Completion of Quick Start screens upon first power up enable measurements, but additional steps may be required to program the analyzer for the desired measurement application. This section covers
the following programming and configuration functions;
1. Selecting measurement type or sensor type (all sections)
2. Identifying the preamp location (pH-see Sec. 6.2)
3. Enabling manual temperature correction and entering a reference temperature (all sections)
4. Enabling sample temperature correction and entering temperature correction slope (selected sections)
5. Defining measurement display resolution (pH)
6. Defining measurement display units (all sections)
7. Adjusting the input filter to control display and output reading variability or noise (all sections)
8. Selecting a measurement range (conductivity – see Sec’s 6.4, 6.5)
9. Entering a cell constant for a contacting sensor (see Sec’s 6.4, 6.5)
10. Entering a temperature element/RTD offset or temperature slope (conductivity-see Sec’s 6.4)
11. Creating an application-specific concentration curve (conductivity-see Sec’s 6.4, 6.5)
To fully configure the analyzer for each installed measurement board, you may use the following:
1. Reset Analyzer function to reset factory defaults and configure the measurement board to the desired
measurement. Follow the Reset Analyzer menu (Fig. 5-5) to reconfigure the analyzer to display new
measurements or measurement units.
2. Program menus to adjust any of the programmable configuration items. Use the following configuration
and programming guidelines for the applicable measurement.
29
MODEL 1057
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
6.2 pH MEASUREMENT PROGRAMMING
6.2.1 Description
This section describes how to configure the Model 1057 analyzer for pH measurements. The following programming and
configuration functions are covered.
TABLE 6-1. pH Measurement Programming
Measure
pH
default setting
Description
Sec.
Menu function:
6.2.2
Measurement type:
6.2.3
Preamp location:
6.2.4
Solution temperature correction Off
Select Off, ultra-pure, high pH, custom
6.2.5
Temp coefficient
Enter the temp coefficient
6.2.6
Resolution:
6.2.7
Filter:
6.2.8
Reference Z:
pH
Analyzer
(custom)
0.01pH
4 sec
Low
Select pH, ORP, Redox, Ammonia, Fluoride, Custom ISE
Identify preamp location
Select 0.01pH or 0.1pH for pH display resolution
Override the default input filter, enter 0-999 seconds
Select low or high reference impedance
A detailed flow diagram for pH programming is provided at the end of Sec. 6 to guide you through
all basic programming and configuration functions.
To configure the pH measurement board:
1. Press MENU
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to
pH. Press ENTER.
The adjacent screen format will appear (factory defaults
are shown). To program any function, scroll to the
desired item and press ENTER.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Configure
Measure:
pH
Preamp:
Analyzer
Sol’n Temp Corr: Off
T Coeff: -0.029pH/°C
Resolution: 0.01pH
Filter:
4 sec
Reference Z:
Low
The following sub-sections provide you with the initial display screen that appears for each configuration function.
Use the flow diagram for pH programming at the end of Sec. 6 and the Model 1057 live screen prompts for each
function to complete configuration and programming.
6.2.2 Measurement
The display screen for selecting the measurement is
shown. The default value is displayed in bold type.
Refer to the pH/ORP Programming flow diagram to
complete this function.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Measurement
pH
ORP
Redox
Ammonia
Fluoride
Custom ISE
6.2.3 Preamp
The display screen for identifying the Preamp location is
shown. The default value is displayed in bold type.
Refer to the pH/ORP Programming flow diagram to
complete this function.
30
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Preamp
Analyzer
Sensor/JBox
MODEL 1057
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
6.2.4 Solution Temperature Correction
The display screen for selecting the Solution
temperature correction algorithm is shown. The default
value is displayed in bold type. Refer to the pH/ORP
Programming flow diagram to complete this function.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Sol’n Temp Corr.
Off
Ultra Pure Water
High pH
Custom
6.2.5 Temperature Coefficient
The display screen for entering the custom solution temperature coefficient is shown. The default value is displayed in bold type.
Refer to the pH/ORP
Programming flow diagram to complete this function.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Sol’n Temp Coeff.
- 0.032pH/ºC
6.2.6 Resolution
The display screen for selecting 0.01pH or 0.1pH for pH
display resolution is shown. The default value is displayed
in bold type. Refer to the pH/ORP Programming flow
diagram to complete this function.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Resolution
0.01pH
0.1pH
6.2.7 Filter
The display screen for entering the input filter value in
seconds is shown. The default value is displayed in
bold type. Refer to the pH/ORP Programming flow diagram
to complete this function.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Input filter
04 sec
6.2.8 Reference Impedence
The display screen for selecting Low or High Reference
impedance is shown. The default value is displayed in
bold type. Refer to the pH/ORP Programming flow diagram
to complete this function.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Reference Z
Low
High
6.3 ORP MEASUREMENT PROGRAMMING
6.3.1 Description
The section describes how to configure the Model 1057 analyzer for ORP measurements. The following programming
and configuration functions are covered:
TABLE 6-2. ORP Measurement Programming
Measure
ORP
Sec.
Menu function:
6.3.2
Measurement type:
6.3.3
Preamp location:
6.3.4
Filter:
6.3.5
Reference Z:
default
pH
Analyzer
4 sec
Low
Description
Select pH, ORP, Redox, Ammonia, Fluoride, Custom ISE
Identify preamp location
Override the default input filter, enter 0-999 seconds
Select low or high reference impedance
31
MODEL 1057
A detailed flow diagram for ORP programming is
provided at the end of Sec. 6 to guide you through
all basic programming and configuration functions.
To configure the ORP measurement board:
1. Press MENU
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to
ORP. Press ENTER.
The adjacent screen format will appear (factory defaults
are shown). To program any displayed function, scroll
to the desired item and press ENTER.
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Configure
Measure:
pH
Preamp: Analyzer
Flter:
4 sec
Reference Z: Low
The following sub-sections provide you with the initial display screen that appears for each configuration function.
Use the flow diagram for ORP programming at the end of Sec. 6 and the Model 1057 live screen prompts for
each function to complete configuration and programming.
6.3.2 Measurement
The display screen for selecting the measurement is
shown. The default value is displayed in bold type.
Refer to the pH/ORP Programming flow diagram to
complete this function.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
6.3.3 Preamp
The display screen for identifying the Preamp location is
shown. The default value is displayed in bold type.
Refer to the pH/ORP Programming flow diagram to
complete this function.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
6.3.4 Filter
The display screen for entering the input filter value in
seconds is shown. The default value is displayed in
bold type. Refer to the pH/ORP Programming flow diagram
to complete this function.
6.3.5 Reference Impedence
The display screen for Selecting Low or high Reference
impedance is shown. The default value is displayed in
bold type. Refer to the pH/ORP Programming flow diagram
to complete this function.
32
1.234ºC
1.234ºC
1.234ºC
SN Measurement
pH
ORP
Redox
Ammonia
Fluoride
Custom ISE
1.234ºC
1.234ºC
1.234ºC
SN Preamp
Analyzer
Sensor/JBox
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Input filter
04 sec
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Reference Z
Low
High
MODEL 1057
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
6.4 CONTACTING CONDUCTIVITY MEASUREMENT PROGRAMMING
6.4.1 Description
The section describes how to configure the Model 1057 analyzer for conductivity measurements using contacting
conductivity sensors. The following programming and configuration functions are covered.
TABLE 6-3. Contacting Conductivity Measurement Programming
Measure
Contacting
Conductivity
default
Description
Sec.
Menu function:
6.4.2
Type:
6.4.3
Measure:
6.4.4
Range:
Auto
6.4.5
Cell K:
1.00000/cm
6.4.6
RTD Offset:
0.00ºC
Enter the RTD Offset
6.4.7
RTD Slope:
0
Enter the RTD Slope
6.4.8
Temp Comp:
6.4.9
Slope:
6.4.10
Ref Temp:
6.4.11
Filter:
6.4.12
Custom
6.4.13
Cal Factor:
2-Electrode
Conductivity
Slope
2.00%/°C
25.0°C
2 sec
Setup
0.95000/cm
Select 2-Electrode or 4-Electrode type sensors
Select Conductivity, Resistivity, TDS. Salinity or % conc
Select measurement Auto-range or specific range
Enter the cell Constant for the sensor
Select Temp Comp: Slope, Neutral Salt, Cation or Raw
Enter the linear temperature coefficient
Enter the Reference temp
Override the default input filter, enter 0-999 seconds
Enter 2-5 data points in ppm and µS/cm for custom curves
Enter the Cal Factor for 4-Electrode sensors from the sensor tag
A detailed flow diagram for contacting conductivity programming is provided at the end of Sec. 6 to
guide you through all basic programming and configuration functions.
To configure the contacting conductivity measurement
board:
1. Press MENU
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to
contacting conductivity. Press ENTER.
The adjacent screen format will appear (factory defaults
are shown). To program any displayed function, scroll
to the desired item and press ENTER.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Configure
Type:
2-Electrode
Measure:
Cond
Range:
Auto
Cell K: 1.00000/cm
RTD Offset: 0.00ºC
RTD Slope:
0
Temp Comp: Slope
Slope:
2.00%/°C
Ref Temp:
25.0°C
Filter:
2 sec
Custom Setup
The following sub-sections provide you with the initial display screen that appears for each configuration function.
Use the flow diagram for contacting conductivity programming at the end of Sec. 6 and the Model 1057 live
screen prompts for each function to complete configuration and programming.
6.4.2 Sensor Type
The display screen for selecting 2-Electrode or
4-Electrode type sensors is shown. The default value
is displayed in bold type. Refer to the contacting
conductivity Programming flow diagram to complete this
function.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Type
2-Electrode
4-Electrode
33
MODEL 1057
6.4.3 Measure
The display screen for selecting the measurement is
shown. The default value is displayed in bold type.
Refer to the contacting conductivity Programming flow
diagram to complete this function.
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Measurement
Conductivity
Resistivity
TDS
Salinity
NaOH (0-12%)
HCl (0-15%)
Low H2SO4
High H2SO4
NaCl (0-20%)
Custom Curve
6.4.4 Range
The display screen for Selecting Auto-ranging or a specific
range is shown. The default value is displayed in bold
type. Note: Ranges are shown as conductance, not
conductivity. Refer to the contacting conductivity
Programming flow diagram to complete this function.
6.4.5 Cell Constant
The display screen for entering a cell Constant for the
sensor is shown. The default value is displayed in bold
type. Refer to the contacting conductivity Programming
flow diagram to complete this function.
6.4.6 RTD Offset
The display screen for Entering the RTD Offset for the
sensor is shown. The default value is displayed in bold
type. Refer to the contacting conductivity Programming
flow diagram to complete this function.
6.4.7 RTD Slope
The display screen for entering the RTD slope for the
sensor is shown. The default value is displayed in bold
type. Refer to the contacting conductivity Programming
flow diagram to complete this function.
6.4.8 Temp Comp
The display screen for Selecting Temperature
Compensation as Slope, Neutral Salt, Cation or Raw is
shown. The default value is displayed in bold type.
Refer to the contacting conductivity Programming flow
diagram to complete this function.
34
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Range
Auto
50 µS
500 µS
2000 µS
20 mS
200 mS
600 mS
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Cell Constant
1.00000 /cm
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN RTD Offset
0.00°C
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN RTD Slope
2.00%/ºC
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Temp Comp
Slope
Neutral Salt
Cation
Raw
MODEL 1057
6.4.9 Slope
The display screen for Entering the conductivity/temp
Slope is shown. The default value is displayed in bold
type. Refer to the contacting conductivity Programming
flow diagram to complete this function.
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
SN Slope
2.00 %/ºC
6.4.10 Reference Temp
The display screen for manually entering the Reference
temperature is shown. The default value is displayed in
bold type. Refer to the contacting conductivity
Programming flow diagram to complete this function.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
6.4.11 Filter
The display screen for entering the input filter value in
seconds is shown. The default value is displayed in
bold type. Refer to the contacting conductivity
Programming flow diagram to complete this function.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
6.4.12 Custom Setup
The display screens for creating a custom curve for
converting conductivity to concentration is shown.
Refer to the contacting conductivity Programming
flow diagram to complete this function.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
When the custom curve data entry is complete, press
ENTER. The display will confirm the determination of a
custom curve fit to the entered data by displaying this
screen:
If the custom curve fit is not completed or is
unsuccessful, the display will read as follows and the
screen will return to the beginning custom curve screen.
6.4.13 Cal Factor
Upon initial installation and power up, if 4-electrode
was selected for the sensor type in the Quick Start
menus, the user enters a Cell Constant and a “Cal
Factor” using the instrument keypad. The cell constant
is needed to convert measured conductance to
conductivity as displayed on the analyzer screen. The
“Cal Factor” entry is needed increase the accuracy of
the live conductivity readings, especially at low conductivity readings below 20uS/cm. Both the Cell Constant
and the “Cal Factor” are printed on the tag attached to
the 4-electrode sensor/cable.
1.234ºC
1.234ºC
1.234ºC
1.234ºC
1.234ºC
1.234ºC
SN Ref Temp
(25.0ºC normal)
+25.0ºC
1.234ºC
1.234ºC
1.234ºC
SN Input filter
02 sec
1.234ºC
1.234ºC
1.234ºC
SN Custom Curve
Configure
Enter Data Points
Calculate Curve
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Calculate Curve
Custom curve
fit completed.
In Process Cal
recommended.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Calculate Curve
Failure
The display screen for entering Cal Factor is shown.
The default value is displayed in bold type. If necessary
after initial installation and start-up, enter the “Cal
Factor” as printed on the sensor tag.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Cal Factor
0.95000/cm
35
FIGURE 6-1 Configuring pH/ORP Measurements
MODEL 1057
36
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
SECTION 6.0
PROGRAMMING THE MEASUREMENTS
FIGURE 6-2 Configure Contacting Measurements
MODEL 1057
37
38
This page left blank intentionally
MODEL 1057
SECTION 7.0
CALIBRATION
SECTION 7.0
CALIBRATION
7.1
7.2
7.3
7.4
7.5
CALIBRATION – INTRODUCTION
pH CALIBRATION
ORP CALIBRATION
CONTACTING CONDUCTIVITY CALIBRATION
TEMPERATURE CALIBRATION
7.1 CALIBRATION – INTRODUCTION
Calibration is the process of adjusting or standardizing
the analyzer to a lab test or a calibrated laboratory
instrument, or standardizing to some known reference
(such as a commercial buffer).
The auto-recognition feature of the analyzer will enable
the appropriate calibration screens to allow calibration for
any single sensor configuration or dual sensor configuration of the analyzer. Completion of Quick Start upon
first power up enables live measurements but does not
ensure accurate readings in the lab or in process.
Calibration should be performed with each attached
sensor to ensure accurate, repeatable readings.
1. Auto buffer cal for pH (pH Cal - Sec.7.2)
2. Manual buffer cal for pH (pH Cal - Sec.7.2)
3. Set calibration stabilization criteria for pH (pH Cal Sec.7.2)
4. Standardization calibration (1-point) for pH, ORP
and Redox (pH Cal - Sec.7.2 and 7.3)
5. Entering the cell constant of a conductivity
sensor (Conductivity Cal - Sec. 7.4 and 7.5)
6. Calibrating the sensor in a conductivity standard
Conductivity Cal - Sec. 7.4 and 7.5)
7. Calibrating the analyzer to a laboratory
instrument (Contacting Conductivity Cal - Sec.7.4)
8. Enter a manual reference temperature for
temperature compensation of the process
measurement
This section covers the following programming and
configuration functions:
7.2 pH CALIBRATION
7.2.1 DESCRIPTION
New sensors must be calibrated before use. Regular recalibration is also necessary. Use auto calibration instead of manual
calibration. Auto calibration avoids common pitfalls and reduces errors. The analyzer recognizes the buffers and uses
temperature-corrected pH values in the calibration. Once the Model 1057 successfully completes the calibration, it calculates and displays the calibration slope and offset. The slope is reported as the slope at 25°C.
THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1057 WITH A pH SENSOR. THE FOLLOWING
CALIBRATION ROUTINES ARE COVERED.
TABLE 7-1
Measure Sec.
pH
7.2.2
pH Calibration Routines
Menu function:
default
Description
Auto Calibration -
pH
2 point buffer calibration with auto buffer recognition
7.2.3
Manual Calibration -
pH
2 point buffer calibration with manual buffer value entry
7.2.4
Entering A Known Slope Value - pH
Slope calibration with manual entry of known slope value
7.2.5
Standardization -
1 point buffer calibration with manual buffer value entry
pH
A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines.
39
MODEL 1057
To calibrate pH:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to
pH. Press ENTER.
4. Select pH. Press ENTER.
SECTION 7.0
CALIBRATION
The following screen will appear. To calibrate pH or
Temperature scroll to the desired item and press
ENTER.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Calibrate?
pH
Temperature
The following sub-sections show the initial display screen that appears for each calibration routine. Use the flow
diagram for pH calibration at the end of Sec. 7 and the live screen prompts to complete calibration.
7.2.2 AUTO CALIBRATION — pH
This screen appears after selecting pH calibration.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN pH Cal
Buffer Cal
Standardize
Slope:
59.16mV/pH
Offset:
600 mV
Note that pH auto calibration criteria can be changed.
The following criteria can be adjusted:
Stabilization time (default 10 sec.)
Stabilization pH value (default 0.02 pH)
Type of Buffer used for AUTO CALIBRATION
(default is Standard, non-commercial buffers).
The following commercial buffer tables are recognized
by the analyzer:
Standard (NIST plus pH7)
DIN 19267
Ingold
Merck
The following screen will appear to allow adjustment of
these criteria:
The following screen will appear if the auto cal is
successful. The screen will return to the pH Buffer Cal
Menu.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Setup
Stable Time:
10 sec
Stable Delta: 0.02 pH
Buffer:
Standard
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN pH Auto Cal
Slope: 59.16 mV/pH
Offset:
60 mV
The following screens may appear if the auto cal is unsuccessful.
1. A High Slope Error will generate this screen display:
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN pH Auto Cal
High Slope Error
Calculated: 62.11 mV/pH
Max: 62.00 mV/pH
Press EXIT
40
MODEL 1057
SECTION 7.0
CALIBRATION
2. A Low Slope Error will generate this screen display:
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN pH Auto Cal
Low Slope Error
Calculated: 39.11mV/pH
Min: 40.00 mV/pH
Press EXIT
3. An Offset Error will generate this screen display:
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN pH Auto Cal
Offset Error
Calculated: 61.22mV
Max:
60.00mV
Press EXIT
7.2.3 MANUAL CALIBRATION — pH
New sensors must be calibrated before use. Regular
recalibration is also necessary. Use manual calibration
if non-standard buffers are being used; otherwise, use
auto calibration. Auto calibration avoids common pitfalls
and reduces errors.
The adjacent appears after selecting Manual pH calibra7.2.4 ENTERING A KNOWN SLOPE VALUE — pH
If the electrode slope is known from other measurements, it can be entered directly in the Model 1057 analyzer. The slope must be entered as the slope at 25°C.
7.2.5 STANDARDIZATION — pH
The pH measured by the Model 1057 analyzer can be
changed to match the reading from a second or referee
instrument. The process of making the two readings
agree is called standardization. During standardization,
the difference between the two pH values is converted
to the equivalent voltage. The voltage, called the reference
offset, is added to all subsequent measured cell voltages
before they are converted to pH. If a standardized sensor
is placed in a buffer solution, the measured pH will differ
from the buffer pH by an amount equivalent to the
standardization offset.
The following screen may appear if ORP Cal is unsuccessful.
An Offset Error will generate this screen display:
If the ORP Cal is successful, the screen will return to the
Cal sub-menu.
tion.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN pH Manual Cal
Buffer 1
Buffer 2
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN pH Slope@25ºC
59.16 mV/pH
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Enter Value
07.00pH
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Standardize
Offset Error
Calculated: 96mV
Max:
60mV
Press EXIT
41
MODEL 1057
SECTION 7.0
CALIBRATION
7.3 ORP CALIBRATION
7.3.1 DESCRIPTION
For process control, it is often important to make the measured ORP agree with the ORP of a standard solution. During
calibration, the measured ORP is made equal to the ORP of a standard solution at a single point.
THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1057 WITH AN ORP SENSOR. THE FOLLOWING CALIBRATION ROUTINE IS COVERED.
TABLE 7-2
Measure
ORP
ORP Calibration Routine
Sec.
7.3.2
Menu function: default
Description
Standardization — ORP 1 point buffer calibration with manual buffer value entry
A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines.
To calibrate ORP:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to
ORP. Press ENTER.
4. Select ORP. Press ENTER.
The following screen will appear. To calibrate ORP or
Temperature, scroll to the desired item and press
ENTER.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Calibrate?
ORP
Temperature
The following sub-sections show the initial display screen that appears for each calibration routine. Use the flow
diagram for ORP calibration at the end of Sec. 7 and the live screen prompts to complete calibration.
7.3.2 STANDARDIZATION — ORP
For process control, it is often important to make the
measured ORP agree with the ORP of a standard
solution. During calibration, the measured ORP is made
equal to the ORP of a standard solution at a single
point. This screen appears after selecting ORP calibration:
Cal sub-menu.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Enter Value
+0600 mV
If the ORP Cal is successful, the screen will return to the
The following screen may appear if ORP Cal is
unsuccessful.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Standardize
Offset Error
Calculated: 61.22mV
Max:
60.00mV
Press EXIT
42
MODEL 1057
SECTION 7.0
CALIBRATION
7.4 CONTACTING CONDUCTIVITY CALIBRATION
7.4.1 DESCRIPTION
PLACING A NEW CONDUCTIVITY SENSOR IN SERVICE
New conductivity sensors rarely need calibration. The cell
constant printed on the label is sufficiently accurate for most
applications.
CALIBRATING AN IN-SERVICE CONDUCTIVITY SENSOR
1. After a conductivity sensor has been in service for a period
of time, recalibration may be necessary. There are three
ways to calibrate a sensor.
a. Use a standard instrument and sensor to measure the
conductivity of the process stream. It is not necessary to
remove the sensor from the process piping. The temperature
correction used by the standard instrument may not exactly
match the temperature correction used by the Model 1057.
To avoid errors, turn off temperature correction in both the
analyzer and the standard instrument.
b. Place the sensor in a solution of known conductivity and
make the analyzer reading match the conductivity of the
standard solution. Use this method if the sensor can be easily
removed from the process piping and a standard is
available. Be careful using standard solutions having
conductivity less than 100 µS/cm. Low conductivity standards
are highly susceptible to atmospheric contamination. Avoid
calibrating sensors with 0.01/cm cell constants against
conductivity standards having conductivity greater than 100
µS/cm. The resistance of these solutions may be too low for
an accurate measurement. Calibrate sensors with 0.01/cm
cell constant using method c.
c. To calibrate a 0.01/cm sensor, check it against a standard
instrument and 0.01/cm sensor while both sensors are
measuring water having a conductivity between 5 and 10
µS/cm. To avoid drift caused by absorption of atmospheric
carbon dioxide, saturate the sample with air before making
the measurements.
To ensure adequate flow past the sensor during calibration,
take the sample downstream from the sensor. For best
results, use a flow-through standard cell. If the process
temperature is much different from ambient, keep
connecting lines short and insulate the flow cell.
THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1057 WITH AN ATTACHED CONTACTING
CONDUCTIVITY SENSOR. THE FOLLOWING CALIBRATION ROUTINES ARE COVERED.
TABLE 7-3
Contacting Conductivity Calibration Routines
default
Description
Measure
Sec.
Menu function:
Contacting
Conductivity
7.4.2
Cell K:
7.4.3
Zero Cal
Zero the analyzer with the sensor attached
7.4.4
In Process Cal
Standardize the sensor to a known conductivity
7.4.5
Meter Cal
Calibrate the analyzer to a lab conductivity instrument
7.4.6
Cal Factor:
1.00000/cm
0.95000/cm
Enter the cell Constant for the sensor
Enter the Cal Factor for 4-Electrode sensors from the sensor tag
A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines
To calibrate contacting conductivity:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
S1: 1.234µS/cm 1.234ºC
3. Select Sensor 1 or Sensor 2 corresponding to
S2: 12.34pH
1.234ºC
contacting conductivity. Press ENTER.
S3: 12.34µS/cm 1.234ºC
4. Select Conductivity. Press ENTER.
SN Calibrate?
The adjacent screen will appear.
To calibrate
Conductivity or Temperature, scroll to the desired item
and press ENTER.
The following sub-sections show the initial display
screen that appears for each calibration routine. Use
the flow diagram for Conductivity calibration at the
end of Sec. 7 and the live screen prompts for each routine to complete calibration.
The adjacent screen appears
Conductivity calibration:
after
selecting
Conductivity
Temperature
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Calibration
Zero Cal
In Process Cal
Meter Cal
Cell K: 1.00000/cm
43
MODEL 1057
7.4.2 ENTERING THE CELL CONSTANT
New conductivity sensors rarely need calibration. The
cell constant printed on the label is sufficiently accurate
for most applications. The cell constant should be
entered:
• When the unit is installed for the first time
• When the probe is replaced
The display screen for entering a cell Constant for the
sensor is shown. The default value is displayed in bold
type.
7.4.3 ZEROING THE INSTRUMENT
This procedure is used to compensate for small offsets
to the conductivity signal that are present even when
there is no conductivity to be measured. This procedure
is affected by the length of extension cable and should
always be repeated if any changes in extension cable or
sensor have been made. Electrically connect the
conductivity probe as it will actually be used and
place the measuring portion of the probe in air. Be
sure the probe is dry.
The adjacent screen will appear after selecting Zero
Cal from the Conductivity Calibration screen:
The adjacent screen will appear if zero Cal is successful.
The screen will return to the conductivity Cal Menu.
SECTION 7.0
CALIBRATION
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Cell Constant
1.00000 /cm
S1: 1.234µS/cm 1.234ºC
S2: 12.34pH
1.234ºC
S3: 12.34µS/cm
1.234ºCSN Zero Cal
In Air
In Water
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Zero Cal
Sensor Zero Done
The adjacent screen may appear if zero Cal is unsuccessful.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Zero Cal
Sensor Zero Fail
Offset too high
Press EXIT
7.4.4 CALIBRATING THE SENSOR IN A CONDUCTIVITY
STANDARD (IN PROCESS CAL)
This procedure is used to calibrate the sensor and
analyzer against a solution of known conductivity.
This is done by submerging the probe in the sample of
known conductivity, then adjusting the displayed value, if
necessary, to correspond to the conductivity value of
the sample. Turn temperature correction off and use the
conductivity of the standard. Use a calibrated thermometer to measure temperature. The probe must be
cleaned before performing this procedure.
The adjacent screen will appear after selecting In
Process Cal from the Conductivity Calibration screen:
44
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN InProcess Cal
Wait for stable
reading.
MODEL 1057
SECTION 7.0
CALIBRATION
The adjacent screen will appear if In Process Cal is successful. The screen will return to the conductivity Cal
Menu.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
The adjacent screen may appear if In Process Cal is
unsuccessful. The screen will return to the conductivity
Cal Men
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN InProcess Cal
Updated cell
constant:
1.00135/cm
1.234ºC
1.234ºC
1.234ºC
SN InProcess Cal
Calibration
Error
Press EXIT
7.4.5 CALIBRATING THE SENSOR TO A LABORATORY
INSTRUMENT (METER CAL)
This procedure is used to check and correct the
conductivity reading of the Model 1057 using a laboratory
conductivity instrument. This is done by submerging the
conductivity probe in a bath and measuring the conductivity of a grab sample of the same bath water with a
separate laboratory instrument. The Model 1057 reading
is then adjusted to match the conductivity reading of the
lab instrument.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
The adjacent screen will appear after selecting Meter
Cal from the Conductivity Calibration screen:
SN Meter Cal
Use precision
resistors only
After pressing ENTER, the display shows the live value
measured by the sensor
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
1.234ºC
1.234ºC
1.234ºC
SN Enter Value
xx.xx kΩ
If the meter cal is successful the screen will return to the
conductivity Cal Menu.
The adjacent screen will appear if Meter Cal is unsuccessful.
The screen will return to the conductivity Cal Menu.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Meter Cal
Calibration
Error
Press EXIT
7.4.6 Cal Factor
Upon initial installation and power up, if 4-electrode
was selected for the sensor type in the Quick Start
menus, the user enters a Cell Constant and a “Cal
Factor” using the instrument keypad. The cell constant
is needed to convert measured conductance to conductivity as displayed on the analyzer screen. The “Cal
Factor” entry is needed increase the accuracy of the live
conductivity readings, especially at low conductivity
readings below 20uS/cm. Both the Cell Constant and
the “Cal Factor” are printed on the tag attached to the
4-electrode sensor/cable.
The display screen for entering Cal Factor is shown.
The default value is displayed in bold type. If necessary after initial installation and start-up, enter the “Cal
Factor” as printed on the sensor tag.
S1: 1.234µS/cm
S2: 12.34pH
S3: 12.34µS/cm
1.234ºC
1.234ºC
1.234ºC
SN Cal Factor
0.95000 /cm
45
MODEL 1057
SECTION 7.0
CALIBRATION
7.5 CALIBRATING TEMPERATURE
7.5.1 DESCRIPTION
Most liquid analytical measurements require temperature compensation (except ORP). The Model 1057 performs
temperature compensation automatically by applying internal temperature correction algorithms. Temperature correction can also be turned off. If temperature correction is off, the Model 1057 uses the manual temperature entered by
the user in all temperature correction calculations.
THIS SECTION DESCRIBES HOW TO CALIBRATE TEMPERATURE IN THE MODEL 1057 ANALYZER. THE
FOLLOWING CALIBRATION ROUTINE IS COVERED.
TABLE 7- 11
Measure
Temperature
Temperature Calibration Routine
Sec.
7.5.2
Calibration function: default value
Calibrate
Description
Enter a manual reference temperature for temperature
compensation of the process measurement
A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines.
To calibrate temperature:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to
the desired measurement. Press ENTER.
4. Select Temperature. Press ENTER.
S1: 1.234µS/cm 1.234ºC
S2: 12.34pH
1.234ºC
S3: 12.34µS/cm 1.234ºC
SN Calibrate
+025.0°C
The adjacent screen will appear.
The following sub-section provides you with the initial
display screen that appears for temperature calibration.
Use the flow diagram for Temp calibration at the
end of Sec. 7 to complete calibration.
7.5.2 CALIBRATION
The adjacent screen will appear during Temperature
Cal.
If the sensor Temperature offset is greater than 5 ºC
from the default value, the following screen will
appear:
You may continue by selecting Yes or suspend this
operation by selecting No.
If the Temp Cal is successful, the screen will return to
the Cal Menu.
Note: To select automatic or manual temp compensation
or to program temperature units as °C or °F, refer to
Sec. 5.3 – Programming Temperature in this manual
46
S1: 1.234µS/cm 1.234ºC
S2: 12.34pH
1.234ºC
S3: 12.34µS/cm 1.234ºC
SN Calibrate
Cal in progress.
Please wait.
S1: 1.234µS/cm 1.234ºC
S2: 12.34pH
1.234ºC
S3: 12.34µS/cm 1.234ºC
SN Temp Offset > 5°C
Continue?
No
Yes
SECTION 7.0
CALIBRATION
FIGURE 7-1 Calibrate pH
MODEL 1057
47
FIGURE 7-2 Calibrate ORP
MODEL 1057
48
SECTION 7.0
CALIBRATION
FIGURE 7-3 Calibrate Contacting Conductivity
MODEL 1057
SECTION 7.0
CALIBRATION
49
FIGURE 7-4 Calibrate Temperature
MODEL 1057
50
SECTION 7.0
CALIBRATION
MODEL 1057
SECTION 8.0
RETURN OF MATERIAL
SECTION 8.0
RETURN OF MATERIAL
8.1 GENERAL
To expedite the repair and return of instruments, proper communication between the customer and the factory is
important. Before returning a product for repair, call 1-949-757-8500 for a Return Materials Authorization (RMA)
number.
8.2 WARRANTY REPAIR
The following is the procedure for returning instruments still under warranty:
1.
Call Rosemount Analytical for authorization.
2.
To verify warranty, supply the factory sales order number or the original purchase order number. In the case
of individual parts or sub-assemblies, the serial number on the unit must be supplied.
3.
Carefully package the materials and enclose your “Letter of Transmittal” (see Warranty). If possible, pack the
materials in the same manner as they were received.
4.
Send the package prepaid to:
Rosemount Analytical
Liquid Division
2400 Barranca Parkway
Irvine, CA 92606
Attn: Factory Repair
RMA No. ____________
IMPORTANT
Please see second section of “Return of
Materials Request” form. Compliance with
the OSHA requirements is mandatory for
the safety of all personnel. MSDS forms
and a certification that the instruments have
been disinfected or detoxified are required.
Mark the package: Returned for Repair
Model No. ____
8.3 NON-WARRANTY REPAIR
The following is the procedure for returning for repair instruments that are no longer under warranty:
1.
Call Rosemount Analytical for authorization.
2.
Supply the purchase order number, and make sure to provide the name and telephone number of the individual to be contacted should additional information be needed.
3.
Do Steps 3 and 4 of Section 17.2.
NOTE
Consult the factory for additional information regarding service or repair.
51
WARRANTY
Seller warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufactured
or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expiration
of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen (18)
months from the date of shipment by Seller, whichever period expires first. Consumables, such as glass electrodes, membranes, liquid junctions, electrolyte, o-rings, catalytic beads, etc., and Services are warranted for a period of 90 days
from the date of shipment or provision.
Products purchased by Seller from a third party for resale to Buyer ("Resale Products") shall carry only the warranty extended by
the original manufacturer. Buyer agrees that Seller has no liability for Resale Products beyond making a reasonable commercial
effort to arrange for procurement and shipping of the Resale Products.
If Buyer discovers any warranty defects and notifies Seller thereof in writing during the applicable warranty period, Seller shall, at
its option, promptly correct any errors that are found by Seller in the firmware or Services, or repair or replace F.O.B. point of manufacture that portion of the Goods or firmware found by Seller to be defective, or refund the purchase price of the defective portion
of the Goods/Services.
All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuitable environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any other
cause not the fault of Seller are not covered by this limited warranty, and shall be at Buyer's expense. Seller shall not be obligated
to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by an
authorized Seller representative. All costs of dismantling, reinstallation and freight and the time and expenses of Seller's personnel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Seller.
Goods repaired and parts replaced during the warranty period shall be in warranty for the remainder of the original warranty period
or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Seller and can be amended only in a
writing signed by an authorized representative of Seller. Except as otherwise expressly provided in the Agreement, THERE ARE
NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESS OR IMPLIED, AS TO MERCHANTABILITY, FITNESS
FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES.
RETURN OF MATERIAL
Material returned for repair, whether in or out of warranty, should be shipped prepaid to:
Emerson Process Management
Liquid Division
2400 Barranca Parkway
Irvine, CA 92606
The shipping container should be marked:
Return for Repair
Model _______________________________
The returned material should be accompanied by a letter of transmittal which should include the following information (make a
copy of the "Return of Materials Request" found on the last page of the Manual and provide the following thereon):
1.
2.
3.
4.
5.
Location type of service, and length of time of service of the device.
Description of the faulty operation of the device and the circumstances of the failure.
Name and telephone number of the person to contact if there are questions about the returned material.
Statement as to whether warranty or non-warranty service is requested.
Complete shipping instructions for return of the material.
Adherence to these procedures will expedite handling of the returned material and will prevent unnecessary additional charges
for inspection and testing to determine the problem with the device.
If the material is returned for out-of-warranty repairs, a purchase order for repairs should be enclosed.
ORDERING INFORMATION
The Model 1057 multi-parameter analyzer supports up to three sensor inputs of pH/ORP and contacting conductivity measurements in any combination thus reducing the cost per loop and saving panel space. All configurations of the analyzer include four 4-20mA current outputs and four fully programmable alarm relays with interval
timers. The instrument supports most industrial, commercial, and municipal applications and is ideal for OEMs,
systems integrators and customers requiring multiple inputs and outputs. All Model 1057 units also includeelectrically isolated inputs, 8 local languages, removable connectors for easy wiring, 4 solid plugs for closure of
openings, and panel mount hardware. All configurations of Model 1057 can be ordered to meet UL approval
requirements. Model 1057 is Class I, Div 2 approved for hazardous area installations by CSA.
Model 1057 Three-Input Intelligent Analyzer
CODE
Power
02
24 VDC with four alarm relays
03
84-265 VAC switching, 50/60 Hz with four alarm relays
CODE
20
22
MEASUREMENT 1
Contacting Conductivity
pH/ORP/ISE
CODE
30
32
38
MEASUREMENT 2
Contacting Conductivity
pH/ORP/ISE
None (not available if -40 or -42 are selected)
CODE
40
42
48
MEASUREMENT 3
Contacting Conductivity
pH/ORP/ISE
None
CODE
UL OPTION
No Selection
UL Approval
UL
ACCESSORIES
PART #
23554-00
23820-00
23820-01
9240048-00
DESCRIPTION
Cable Gland Kit - Qty 5 per pack
2" Pipe Mounting Bracket includes U-bolts, mounting bracket, nuts, washers, and
screws (complete). Also accommodates wall / surface mounting.
2" Pipe Mounting Bracket, Stainless Steel
S.S. Tag (specify marking)
53
The right people,
the right answers,
right now.
ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
http://www.raihome.com
Credit Cards for U.S. Purchases Only.
Emerson Process Management
Rosemount Analytical Inc.
2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 757-8500
Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2009