Welker Manual Insertion Probe IOM
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Installation, Operation, and
Maintenance Manual
Welker® Manual Insertion Probe
Models
MIP
The information in this manual has been carefully checked for accuracy and is intended to be used as a
guide for the installation, operation, and maintenance of the Welker equipment described above. Correct
operating and/or installation techniques, however, are the responsibility of the end user. Welker reserves
the right to make changes to this and all products in order to improve performance and reliability.
13839 West Bellfort
Sugar Land, TX 77498-1671
(281) 491-2331 - Office
(800) 776-7267 - USA Only
(281) 491-8344 - Fax
http://www.welkereng.com
© Copyright 2008, Welker, Inc.
TABLE OF CONTENTS
3
1.
GENERAL
1.1
1.2
1.3
1.4
1.5
INTRODUCTION
DESCRIPTION OF PRODUCT
SPECIFICATIONS
OPTIONS
SYSTEM DIAGRAM
2.
INSTALLATION & OPERATIONS
2.1
2.2
2.3
GENERAL
INSTALLING THE PROBE
RETRACTING THE PROBE
3.
MAINTENANCE
3.1
3.2
GENERAL
DISASSEMBLY & MAINTENANCE
3
3
4
5
5
6
6
6
7
8
8
9
Welker®, Welker Jet®, and WelkerScope® are Registered Trademarks owned by Welker, Inc.
IOM-100
Welker® Manual Insertion Probe
Rev: B
Last updated: 9/8/2009
Page 2 of 10
SPECIFICATIONS
1.
GENERAL
1.1
INTRODUCTION
We appreciate your business and your choice of Welker products. The installation,
operation, and maintenance liability for this product becomes that of the purchaser at the
time of receipt. Reading the applicable Installation, Operation, and Maintenance (IOM)
Manual prior to installation and operation of this equipment is required for a full
understanding of its application and performance prior to use.*
If you have any questions, please call 1-800-776-7267 in the USA or 1-281-491-2331.
Notes, Warnings, and Cautions
Notes emphasize information or set it off from the surrounding text.
Caution messages appear before procedures that, if not observed, could result
in damage to equipment.
Warnings alert users to a specific procedure or practice that, if
not followed correctly, could cause personal injury.
*The following procedures have been written for use with standard Welker parts and equipment.
Assemblies that have been modified may have additional requirements and specifications that are not listed
in this manual.*
1.2 DESCRIPTION OF PRODUCT
The Welker Manual Insertion Probe can be manually
inserted and withdrawn from a pipeline through a full
ported ball valve, where the pipeline pressures do not
exceed maximum insertion pressure for the shaft
diameter. When the pressures exceed maximum psi,
we recommend the pipeline being depressurized
before insertion or withdrawal of the Welker Manual
Insertion Probe.
IOM-100
Welker® Manual Insertion Probe
Rev: B
Last updated: 9/8/2009
Page 3 of 10
SPECIFICATIONS
1.3
SPECIFICATIONS
Table 1
Welker Manual Insertion Probe Specifications
Materials of Construction:
Lubricator Body, Carbon Steel; Wetted Parts, 316SS, Lubricator Body, 304LSS;
Wetted Parts, Parts 304LSS
Viscosity Range:
10-◦ 50◦ API Gravity
Temperature Range:
◦
F, Determined by flange rating
Maximum Line Pressure:
PSI; Determined by flange rating
Pipeline Connection:
ANSI, Determined by flange rating
Area Classification:
Can be used in hazardous areas
Insertion/Retraction
Pressure
PSI, Determined by probe shaft diameter
Materials of Construction:
Lubricator Body, Carbon Steel; Wetted Parts, 316SS, Lubricator Body, 304LSS;
Wetted Parts, Parts 304LSS
Table 2
Maximum Insertion/Retraction Pressure
Diameter of Probe Shaft
Maximum Allowable Pressure
3
455 psi
1
255 psi
5
163 psi
3
113 psi
1 inch
64 psi
11/4 inch
40 psi
11/2 inch
28 psi
13/4 inch
20 psi
2 inches
15 psi
/8 inch
/2 inch
/8 inch
/4 inch
IOM-100
Welker® Manual Insertion Probe
Rev: B
Last updated: 9/8/2009
Page 4 of 10
SPECIFICATIONS
1.4
OPTIONS
•
•
•
•
•
•
1.5
Materials of Construction
Pipeline/Isolation Connection
Flow Rates/Pressure Drops/Injection Rates
Inlet/Outlet Ports (Sizes)
High or low temperatures
Higher Pressures and Flange Ratings
SYSTEM DIAGRAM
Figure 1
Probe Diagram
Refer to this Figure throughout the installation, operation, & maintenance process.
IOM-100
Welker® Manual Insertion Probe
Rev: B
Last updated: 9/8/2009
Page 5 of 10
INSTALLATION & OPERATIONS
2.
INSTALLATION & OPERATION INSTRUCTIONS
2.1
GENERAL
After unpacking the unit, check it for compliance and for any damages that may have occurred during
shipment.
Claims for damages caused during shipping must be initiated by the receiver and directed
to the shipping carrier. Welker is not responsible for any damages caused from
mishandling by the shipping company.
When sealing fittings with PTFE tape, refer to the proper sealing instructions for the tape
used.
2.2
INSTALLATION THE PROBE (Refer to Figure 1 on page 5)
1. If pipeline pressure is not below the maximum allowable pressure for installation/retraction of the
probe located in Table 2 of Section 1.3, depressurize the pipeline.
2. Install an open and close valve on the outlet port of the probe.
3. Close all valves on the manual insertion probe.
4. Determine the insertion depth of the probe by measuring from the top of the pipeline isolation valve
to the desired depth in the pipeline.
5. Measuring from the top of the lubricator body to the bottom of the lock collar, move the lock collar
to a point on the probe that is equal to the desired insertion depth, and then tighten the lock collar
firmly.
Example: If the insertion travel desired is 12 inches, measure upward from the top of the lubricator
body to a point on the probe 12 inches high and tighten the lock collar at this position. Therefore, the
travel will be limited to this distance.
This procedure assures the insertion will be limited by the lock collar.
6.
7.
8.
9.
The unit is now ready to be attached to a full ported pipeline isolation valve.
Attach the probe to the pipeline isolation valve.
Slowly open the pipeline isolation valve and check for leaks.
Using the insertion handles on the lock collar, manually push and guide the probe into the pipeline.
Always insert probe slowly, and control the movement of the shaft, this will prevent
the probe from slamming back and forth and possibly injuring someone.
10. Secure the probe into place by guiding the lock collar over the lubricator body stud bolts and tighten
the nuts firmly.
Do not let go of the injection probe until this step is complete!
IOM-100
Welker® Manual Insertion Probe
Rev: B
Last updated: 9/8/2009
Page 6 of 10
INSTALLATION & OPERATIONS
11. Make sure all valves are closed, and check for leaks.
12. The installation process is now complete.
13. If desired, you may attach tubing to the outlet valve of the probe in order to install end-user device.
2.3 RETRACTING THE PROBE (Refer to Figure 1 on page 5)
1.
Close the probe outlet valve, and remove any attached tubing.
If the pipeline pressure is greater than the maximum allowable pressure for
installation/retraction of the probe located in Table 2 of Section 1.3, the pipeline pressure must
be dropped or depressurized.
AT PRESSURES ABOVE MAXIMUM (SEE TABLE 2), YOU WILL NOT BE ABLE
TO CONTROL THE RATE AT THAT THE INSERTION PROBE COMES OUT.
PLEASE RELIEVE THE LINE PRESSURE TO BELOW MAXIMUM ALLOWABLE
PRESSURE (Table 2) BEFORE REMOVING THE PROBE TO AVOID POSSIBLE
INJURY.
2.
You will need to apply a force down on the insertion probe just enough to control the rate at that it is
withdrawn. The pipeline pressure will push the probe out of the pipeline.
i. Push in on the probe handles firmly and remove the stud nuts that hold the lock collar
down.
ii. With the nuts removed, slowly let off the force holding the probe down and allow
pipeline pressure to retract the probe.
iii. If there is no pressure on the pipeline, the insertion probe must be pulled by the handles
through the lubricator body and spool manually.
3.
When the insertion probe has been completely retracted from the pipeline, close the pipeline isolation
valve.
4.
Bleed any trapped pressure in the insertion probe assembly by slowly opening the lubricator body
purge valve.
5.
The insertion probe can now be removed from the pipeline isolation valve.
IOM-100
Welker® Manual Insertion Probe
Rev: B
Last updated: 9/8/2009
Page 7 of 10
MAINTENANCE
3.
MAINTENANCE
3.1
GENERAL
Prior to maintenance or disassembly of the unit, it is advisable to have a repair kit handy
for the system in case of unexpected wear or faulty seals. All maintenance and cleaning
of the unit should be done on a smooth, clean surface.
Under normal operating conditions, in that the probe remains installed in the pipeline, the
probe should not need maintenance unless a leak is apparent, or upon removal of the probe.
If the probe remains permanently installed, it should be serviced at least once every ten years.
In the case of severe service, dirty conditions, excessive retraction and installation, or other
unique applications that may subject the equipment to unpredictable circumstances, a more
frequent maintenance schedule may be appropriate.
Recommended Tools
It would be advisable to have the following tools available for installation of the unit.
However, tools used will vary depending on cylinder model and connectors used.
•
•
(2) adjustable wrenches
600-grit sandpaper
New seals supplied in spare parts kits are not lubricated. They should be lightly coated with
lubrication grease (silicone grease or other) before they are installed into the equipment.
This helps in the installation of the seals while reducing the risk of damage when positioning
them on the parts. After the seals are installed, some additional lubrication can be applied
to shafts or cylinder inner diameters to allow smooth transition of parts.
IOM-100
Welker® Manual Insertion Probe
Rev: B
Last updated: 9/8/2009
Page 8 of 10
MAINTENANCE
3.2
DISASSEMBLY & MAINTENANCE INSTRUCTIONS
1.
2.
3.
4.
5.
6.
7.
8.
9.
Remove the probe from the pipeline following Section 2.3 guidelines.
Place the probe assembly on its side in a clean work place.
Remove the two hex nuts (Part 1) from the two lock down studs (Part 4).
Remove the probe outlet valve.
Push the probe shaft out of the lubricator body.
Replace the O-ring (Part 6), back-up (Part 10), and wiper (Part 12).
Check the surface of the shaft for scratches, and repair as necessary.
Reinstall the probe shaft.
Reinsert the insertion probe to the correct depth, and set the lock collar on the lock
down studs (Part 3).
10. Replace and tighten hex nuts (Part 1) on the lock down studs (Part 4).
11. The probe is now ready to be reinstalled. Please refer to the installation instructions
beginning on page 6 for complete details.
IOM-100
Welker® Manual Insertion Probe
Rev: B
Last updated: 9/8/2009
Page 9 of 10
13839 West Bellfort, Sugar Land, Texas 77498-1671
Phone: (281) 491-2331
Fax: (281) 491-8344
Toll Free: (800) 776-7267
Web Page: www.welkereng.com