Yokogawa ISC450 – Calibration In Place
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
ISC450 – Calibration In Place
There are three choices for calibration of the ISC40 sensor and the ISC450 analyzer. The
simplest, is to select cell constant (automatic). However, there is less then one inch
around the sensor to the pipe an inline process calibration can be done. Alternatively
the instrument can be calibrated using the process solution measured with a standard
instrument. Care must be taken to make the measurement at the reference
temperature since differences in the type of temperature compensation of the
instruments may cause an error. With this method the sensor is not removed from the
process. This method is the most convenient method of calibrating the ISC450G
converter. Since the sensor is immersed in the process, errors caused by installation
characteristics are compensated for.
This Tech Note is designed to guide you through the steps required for in process
calibration.
1.) At the opening Screen – Press the
WRENCH Key
2.) Press the Calibration diamond
Sales News: TNA0909_Rev2
Date: March 3, 2011
3.) Press the Sample diamond
4.) Press the SC1 diamond
5.) Press the Take Sample diamond
Sales News: TNA0909_Rev2
Date: March 3, 2011
6.) While the sample is taking the reading it will
flash “Checking Stability”
7.) Once the reading is stabilize you will see the
following screen. Press the HOME icon, to end
calibration and return to normal measuring
mode.
If and when the cell has been subjected to abrasion (erosion or coating) calibration
may be necessary. The screens will look a little different and there is an additional step.
Sales News: TNA0909_Rev2
Date: March 3, 2011
1.) At the opening Screen – Press the
WRENCH Key
2.) Press the Calibration diamond
3.) Press the Sample diamond
4.) Press the SC1 diamond
Sales News: TNA0909_Rev2
Date: March 3, 2011
5.) This time the screen looks different
because it shows the stored data from the last
calibration. Press the Start Calibration diamond
6.) This will generate a screen that will prompt you
to specify what the conductivity value of the
sample is. Use the numeric key pad to enter the
valve you press Enter.
7.) If the cell constant did not change the
following Message will appear.
Press the HOME icon, to end calibration and
return to normal measuring mode.
Sales News: TNA0909_Rev2
Date: March 3, 2011
8.) If the cell constant has changed the following
screen will appear. Press the Accept Data
diamond, to end calibration. It will take you back
to the sample screen.
9.) Press the HOME icon, to return to normal
measuring mode.
Sales News: TNA0909_Rev2
Date: March 3, 2011