Endress + Hauser Operating Instructions Condumax CLS12/CLS13 Conductivity sensors for high-temperature applications
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
BA01641C/07/EN/01.16
71340911
Products
Solutions
Operating Instructions
Condumax CLS12/CLS13
Conductivity sensors for high-temperature
applications
0
Services
Table of contents
Table of contents
1
Document information . . . . . . . . . 3
1.1
1.2
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2
Basic safety instructions . . . . . . . 4
2.1
2.2
2.3
2.4
2.5
Requirements for personnel . . . . . . . . . . .
Designated use . . . . . . . . . . . . . . . . . . . . . .
Occupational safety . . . . . . . . . . . . . . . . . .
Operational safety . . . . . . . . . . . . . . . . . . .
Product safety . . . . . . . . . . . . . . . . . . . . . . .
3
Incoming acceptance and
product identification . . . . . . . . . . 7
3.1
3.2
3.3
3.4
Incoming acceptance . . . . . . . . . . . . . . . . .
Product identification . . . . . . . . . . . . . . . . .
Scope of delivery . . . . . . . . . . . . . . . . . . . . .
Certificates and approvals . . . . . . . . . . . . .
4
Installation . . . . . . . . . . . . . . . . . . . . . . 9
4.1
4.2
Mounting the sensor . . . . . . . . . . . . . . . . . 9
Post-installation check . . . . . . . . . . . . . . . . 9
4
4
4
4
5
7
7
8
8
5
Electrical connection . . . . . . . . . . . 9
5.1
5.2
5.3
5.4
Connection conditions . . . . . . . . . . . . . . .
Connecting the sensor . . . . . . . . . . . . . . .
Ensuring the degree of protection . . . . .
Post-connection check . . . . . . . . . . . . . . .
6
Commissioning . . . . . . . . . . . . . . . . 12
7
Maintenance . . . . . . . . . . . . . . . . . . 12
8
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . 13
8.1
8.2
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
9
Technical data . . . . . . . . . . . . . . . . . 14
10
EU Declaration of
conformity . . . . . . . . . . . . . . . . . . . . . 17
10
11
11
11
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2
Endress+Hauser
Condumax CLS12/CLS13
Document information
1
Document information
1.1
Warnings
Structure of information
LDANGER
Causes (/consequences)
Consequences of non-compliance (if
applicable)
‣ Corrective action
LWARNING
Causes (/consequences)
Consequences of non-compliance (if
applicable)
‣ Corrective action
LCAUTION
Causes (/consequences)
Consequences of non-compliance (if
applicable)
‣ Corrective action
NOTICE
Meaning
This symbol alerts you to a dangerous situation.
Failure to avoid the dangerous situation will result in a fatal or serious injury.
This symbol alerts you to a dangerous situation.
Failure to avoid the dangerous situation can result in a fatal or serious injury.
This symbol alerts you to a dangerous situation.
Failure to avoid this situation can result in minor or more serious injuries.
This symbol alerts you to situations which may result in damage to property.
Cause/situation
Consequences of non-compliance (if
applicable)
‣ Action/note
1.2
Symbols
Symbol
Meaning
Additional information, tips
Permitted or recommended
Not permitted or not recommended
Reference to device documentation
Reference to page
Reference to graphic
Endress+Hauser
Result of a step
3
Basic safety instructions
2
Basic safety instructions
2.1
Requirements for personnel
Condumax CLS12/CLS13
• Installation, commissioning, operation and maintenance of the measuring system may be
carried out only by specially trained technical personnel.
• The technical personnel must be authorized by the plant operator to carry out the specified
activities.
• The electrical connection may be performed only by an electrical technician.
• The technical personnel must have read and understood these Operating Instructions and
must follow the instructions contained therein.
• Measuring point faults may be repaired only by authorized and specially trained personnel.
Repairs not described in the Operating Instructions provided may only be carried out
directly by the manufacturer or by the service organization.
2.2
Designated use
The sensor measures conductivity in water or steam circuits, such as in the power plant and
energy industry:
• Condensate monitoring
• Boiler feedwater monitoring
• Boiler blowdown monitoring
The sensor can be used in all applications with a low conductivity and high, CLS13 with very
high, temperatures and pressures.
Sensor versions with explosion protection approval according to ATEX, FM or CSA are suitable
for use in hazardous areas.
Use of the device for any purpose other than that described, poses a threat to the safety of
people and of the entire measuring system and is therefore not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.
2.3
Occupational safety
As the user, you are responsible for complying with the following safety conditions:
• Installation guidelines
• Local standards and regulations
• Regulations for explosion protection
Electromagnetic compatibility
• The product has been tested for electromagnetic compatibility in accordance with the
applicable European standards for industrial applications.
• The electromagnetic compatibility indicated applies only to a product that has been
connected in accordance with these Operating Instructions.
2.4
1.
4
Operational safety
Before commissioning the entire measuring point, verify that all connections are
correct. Ensure that electrical cables and hose connections are undamaged.
Endress+Hauser
Condumax CLS12/CLS13
Basic safety instructions
2.
Do not operate damaged products, and safeguard them to ensure that they are not
operated inadvertently. Label the damaged product as defective.
3.
If faults cannot be rectified:
Take the products out of operation and safeguard them to ensure that they are not
operated inadvertently.
2.5
Product safety
2.5.1
State of the art
The product is designed to meet state-of-the-art safety requirements, has been tested, and
left the factory in a condition in which it is safe to operate. The relevant regulations and
European standards have been observed.
2.5.2
Electrical equipment in hazardous areas
• The sensors CLS12 / CLS13 have been developed and manufactured in compliance with
applicable European standards and guidelines and are suitable for use in hazardous areas.
The EC type examination certificate confirms compliance with the harmonized European
standards for using the sensors in hazardous areas. The corresponding EU declaration of
conformity is part of this document.
• The sensors may only be operated on suitable intrinsically safe circuits. Make sure that the
maximal permissible sensor input characteristic values, the maximum permissible
inductance Li and capacitance values Ci in these circuits and the ambient temperature
ranges indicated are not exceeded.
• The electrical connection must be made according to the wiring diagram of the transmitter.
• Metallic process connection parts must be mounted at the mounting location
electrostatically conductive (< 1 MΩ).
• The maximum permissible cable length is limited by the maximum permissible
characteristic values of the transmitter: the total of the maximum permissible inductance Li
and capacitance values Ci for the sensor and measuring cable may not exceed the maximum
permissible inductance Lo and capacitance values Co for the transmitter.
• The sensors CLS12 and CLS13 must be installed in such a way that they are protected
against friction and impact.
• Full compliance with regulations for electrical systems in hazardous areas (e.g. EN/IEC
60079-14) is mandatory when using the devices and sensors.
Endress+Hauser
5
Basic safety instructions
Condumax CLS12/CLS13
Temperature classes
Name
Type
Medium temp. Ta for temperature class (Tn)
Cat.
x1 x2 x3 x4
Condumax
CLS12
-
*
** *
A
-20 °C ≤ Ta ≤ +160 °C (T3)
-20 °C ≤ Ta ≤ +125 °C (T4)
-20 °C ≤ Ta ≤ +75 °C (T6)
II 1G
Condumax
CLS13
-
*
** *
A
-20 °C ≤ Ta ≤ +250 °C (T2)
-20 °C ≤ Ta ≤ +190 °C (T3)
-20 °C ≤ Ta ≤ +125 °C (T4)
-20 °C ≤ Ta ≤ +75 °C (T6)
II 1G
x1 ... Measuring range and cell constant (no Ex relevance)
x2 ... Process connection / material (no Ex relevance)
x3 ... Cable entry (no Ex relevance)
x4 ... Temperature sensor: A = Pt 100
If the specified medium temperatures are complied with, temperatures that are not permitted
for the respective temperature class will not occur on the equipment.
The following connection values are safety limits which may not be exceeded when connecting
to the transmitter:
Parameters
Connection data
Supply circuit
Intrinsically safe
Maximum input voltage Ui
15 V
Maximum input current Ii
30 mA
Maximum input power Pi
130 mW
Maximum internal capacitance Ci
Negligible
Maximum internal inductance Li
Negligible
Measuring cable CYK71
Maximum internal capacitance Ci
1 nF/m
Maximum internal inductance Li
6 µH/m
FM/CSA IS/NI Cl.1 Div.1&2 Gr. A-D
Observe the documentation and the control drawings of the transmitter.
6
Endress+Hauser
Condumax CLS12/CLS13
Incoming acceptance and product identification
3
Incoming acceptance and product identification
3.1
Incoming acceptance
1.
Verify that the packaging is undamaged.
Notify your supplier of any damage to the packaging.
Keep the damaged packaging until the matter has been settled.
2.
Verify that the contents are undamaged.
Notify your supplier of any damage to the delivery contents.
Keep the damaged products until the matter has been settled.
3.
Check the delivery for completeness.
Check it against the delivery papers and your order.
4.
Pack the product for storage and transportation in such a way that it is protected
against impact and moisture.
The original packaging offers the best protection.
The permitted ambient conditions must be observed (see "Technical data").
If you have any questions, please contact your supplier or your local sales center.
3.2
Product identification
3.2.1
Nameplate
The nameplate provides you with the following information on your device:
• Manufacturer identification
• Extended order code
• Serial number
• Safety information and warnings
• Cell constant (nominal value)
• Protection class
• Ex labeling on hazardous area versions
‣ Compare the data on the nameplate with your order.
3.2.2
Product identification
Interpreting the order code
The order code and serial number of your product can be found in the following locations:
• On the nameplate
• In the delivery papers
Obtaining information on the product
1.
Go to the product page for your product on the Internet.
2.
At the bottom of the page, select the "Online Tools" link followed by "Check your device
features".
An additional window opens.
Endress+Hauser
7
Incoming acceptance and product identification
3.
Condumax CLS12/CLS13
Enter the order code from the nameplate into the search field, and then select "Show
details".
You will receive information on each feature (selected option) of the order code.
Manufacturer's address
Endress+Hauser Conducta GmbH+Co. KG
Dieselstraße 24
D-70839 Gerlingen
3.3
Scope of delivery
The scope of delivery includes:
• Sensor in the version ordered
• Operating Instructions
3.4
Certificates and approvals
3.4.1
mark
Declaration of conformity
The product meets the requirements of the harmonized European standards. As such, it
complies with the legal specifications of the EU directives. The manufacturer confirms
successful testing of the product by affixing to it the mark.
3.4.2
Hazardous area approvals
ATEX II 1G Ex ia IIC T3/T4/T6 Ga (CLS12)
ATEX II 1G Ex ia IIC T2/T3/T4/T6 Ga (CLS13)
In conjunction with Liquiline M CM42 transmitter:
FM/CSA IS/NI Cl.I, Div.1&2 Gr. A-D T6 Ta
Cl.I Zone 0 AEx ia IIC T6 Ta
CI.I Zone 2 IIC T6 Ta
3.4.3
Manufacturer inspection certificate
Stating the individual cell constant
3.4.4
Ex-certification body
TÜV Rheinland Industrie Service GmbH
Cologne
8
Endress+Hauser
Condumax CLS12/CLS13
4
Installation
Installation
4.1
Mounting the sensor
‣ Mount the sensor directly on a process connection that matches the version. Alternatively
you can install the sensor in a flow assembly CLA751.
Ensure that the electrodes are fully immersed in the medium during measurement. If the
sensor is being used in the ultrapure water range, you must work under air-evacuated
conditions. Otherwise the CO2 in the air can dissolve in the water and its (weak)
dissociation can increase the conductivity by up to 3 µS/cm.
4.2
Post-installation check
• Are the sensor and cable undamaged?
• Is the sensor installed in the process connection and is not suspended from the cable?
5
Electrical connection
LWARNING
Device is live
Incorrect connection may result in injury or death
‣ The electrical connection may be performed only by an electrical technician.
‣ The electrical technician must have read and understood these Operating Instructions and
must follow the instructions contained therein.
‣ Prior to commencing connection work, ensure that no voltage is present on any cable.
Endress+Hauser
9
Electrical connection
Condumax CLS12/CLS13
5.1
Connection conditions
5.1.1
Connection at a glance
Zone 1
IIC 2G
EX
Ex ia Ga
Ex ia Ga
CPS41
CPS42
CPS441
CPS11
CPS12
CPS71
CPS72
CPS91
CPS13
CPS64
CPS471
CPS491
Zone 0
IIC 1G
Zone 2
IIC 3G
EX
CLS50 CLS54
CLS12 CLS13
CLS15 CLS16 CLS21
A0031175
1
10
Electrical connection in hazardous areas
Endress+Hauser
Condumax CLS12/CLS13
5.2
Electrical connection
Connecting the sensor
The sensor is connected via the CYK71 measuring cable with a shield. The wiring diagram is
provided in the Operating Instructions of the transmitter used.
}
GN
WH
YE
BN n.c.
3
}
1
Pt 100
GN/YE
Outer shield
(see wiring diagram
of transmitter)
Sensor
Plug
) shield (outer electrode)
2 conductivity (inner electrode)
Cable
GY
Coax BK
Coax inner
A0024205-EN
2
Measuring cable CYK71
A VMB junction box and another CYK71 cable are required for the cable extension.
5.3
Ensuring the degree of protection
Only the mechanical and electrical connections which are described in these instructions and
which are necessary for the required, designated use, may be carried out on the device
delivered.
‣ Exercise care when carrying out the work.
Otherwise, the individual types of protection (Ingress Protection (IP), electrical safety, EMC
interference immunity) agreed for this product can no longer be guaranteed due, for example,
to covers being left off or cable (ends) which are loose or insufficiently secured.
5.4
Post-connection check
Device condition and specifications
Notes
Are the outside of the sensor, assembly, cable undamaged?
Visual inspection
Electrical connection
Notes
Are the installed cables strain-relieved and not twisted?
Is a sufficient length of the cable cores stripped, and is it positioned
in the terminal correctly?
Check the fit (by pulling gently)
Are all the screws terminals properly tightened?
Tighten
Are all cable entries mounted, tightened and leak-tight?
For lateral cable entries, make sure the cables
loop downwards to allow water to drip off
Are all cable entries installed downwards or mounted laterally?
Endress+Hauser
11
Commissioning
6
Condumax CLS12/CLS13
Commissioning
Before first commissioning, check if:
• the sensor is correctly installed
• the electrical connection is correct.
If using an assembly with automatic cleaning, check that the cleaning medium (e.g. water or
air) is connected correctly.
LWARNING
Escaping process medium
Risk of injury from high pressure, high temperatures or chemical hazards
‣ Before applying compressed air to an assembly with cleaning facility, make sure the
connections are correctly fitted.
‣ Do not install the assembly in the process if you cannot make the correct connection
reliably.
7
Maintenance
LCAUTION
Corrosive chemicals
Danger of chemical burns to the eyes and skin. Danger of damage to clothing and equipment
‣ It is absolutely essential to protect the eyes and hands properly when working with acids,
bases and organic solvents!
‣ Wear protective goggles and safety gloves.
‣ Clean away splashes on clothes and other objects to prevent any damage.
‣ Pay particular attention to the information provided in the safety data sheets for the
chemicals used.
LWARNING
Hydrofluoric acid and mineral acids
Risk of serious or fatal injury from caustic burns
‣ Wear protective goggles to protect your eyes.
‣ Wear protective gloves and appropriate protective clothing.
‣ Avoid all contact with the eyes, mouth and skin.
‣ If using hydrofluoric acid, only use plastic vessels.
LWARNING
Thiocarbamide
Harmful if swallowed. Limited evidence of carcinogenicity. Possible risk of harm to the unborn
child. Dangerous for the environment with long-term effects.
‣ Wear protective goggles, protective gloves and appropriate protective clothing.
‣ Avoid all contact with the eyes, mouth and skin.
‣ Avoid releases into the environment.
12
Endress+Hauser
Condumax CLS12/CLS13
Repair
Clean away fouling on the sensor as follows depending on the type of fouling:
1.
Oily and greasy films:
Clean with grease remover, e.g. alcohol, as well as hot water and (alkaline) agents
containing surfactants (e.g. dishwashing detergent).
2.
Lime, cyanide and metal hydroxide buildup and low solubility organic buildup:
Dissolve buildup with diluted hydrochloric acid (3 %) and then rinse thoroughly with
plenty of clear water.
3.
Sulfidic buildup (from flue gas desulfurization or sewage treatment plants):
Use a mixture of hydrochloric acid (3 %) and thiocarbamide (commercially available)
and then rinse thoroughly with plenty of clear water.
4.
Buildup containing proteins (e.g. food industry):
Use a mixture of hydrochloric acid (0.5 %) and pepsin (commercially available) and then
rinse thoroughly with plenty of clear water.
5.
Readily soluble biological buildup:
Rinse with pressurized water.
After cleaning or regeneration, you must rinse the sensor thoroughly with water .
8
Repair
8.1
Return
The product must be returned if repairs or a factory calibration are required, or if the wrong
product was ordered or delivered. As an ISO-certified company and also due to legal
regulations, Endress+Hauser is obliged to follow certain procedures when handling any
returned products that have been in contact with medium.
To ensure swift, safe and professional device returns, please read the return procedures and
conditions at www.endress.com/support/return-material.
8.2
Disposal
The device contains electronic components and must therefore be disposed of in accordance
with regulations on the disposal of electronic waste.
Observe the local regulations.
Endress+Hauser
13
Technical data
9
Technical data
9.1
Input
9.1.1
Measured values
Condumax CLS12/CLS13
• Conductivity
• Temperature
9.1.2
Measuring ranges
Conductivity
CLS12 / CLS13 -A
CLS12 / CLS13 -B
Temperature
CLS12
CLS13
9.1.3
-20 to 160 °C (-4 to 320 °F)
-20 to 250 °C (-4 to 480 °F)
Cell constant
CLS12 / CLS13 -A
CLS12 / CLS13 -B
9.1.4
(in relation to water at 25 °C (77 °F))
0.04 to 20 μS/cm
0.10 to 200 μS/cm
k = 0.01 cm–1
k = 0.1 cm–1
Temperature compensation
Pt 100 (class B as per IEC 60751)
9.2
Environment
9.2.1
Ambient temperature range
-20 to +60 °C (-4 to 140 °F)
9.2.2
Storage temperature
-25 to +80 °C (-10 to +180 °F)
9.2.3
Degree of protection
IP67
14
Endress+Hauser
Condumax CLS12/CLS13
Technical data
9.3
Process
9.3.1
Process temperature
CLS12
-20 to 160 ˚C (-4 to 320 ˚F)
CLS13
-20 to 250 °C (-4 to 480 °F)
9.3.2
Process pressure
CLS12
Without flow assembly CLA751
Up to 100 °C (212 °F):
1 to 41 bar (15 to 595 psi), absolute
Up to 160 °C (320 °F):
1 to 7 bar (15 to 102 psi), absolute
In flow assembly CLA751
1 to 13 bar (15 to 185 psi), absolute
CLS13
1 to 41 bar (15 to 595 psi), absolute
9.3.3
Temperature/pressure ratings
CLS12
p (abs.)
[psi] [bar]
595
41
435
30
290
20
102
7
-20
0
-4
32
30
70
100
160
212
320
[°C]
T
[°F]
A0032120-EN
3
Mechanical pressure-temperature resistance
Endress+Hauser
15
Technical data
Condumax CLS12/CLS13
CLS13
p (abs.)
[psi] [bar]
595
41
435
30
290
20
145
10
-20
20
-4
68
60
100 140
180 220 250
212
356
482
[°C]
T
[°F]
A0032124-EN
4
Mechanical pressure-temperature resistance
9.4
Mechanical construction
9.4.1
Weight
CLS12
Approx. 1.4 kg (3.1 lbs) depending on version
CLS13
Approx. 1.7 kg (3.75 lbs) depending on version
9.4.2
Materials
CLS12
Electrodes
Process connection
Terminal head
Insulator body
Stainless steel 1.4571 (AISI 316Ti)
Stainless steel 1.4571 (AISI 316Ti)
Die-cast aluminum
EPDM, PEEK
CLS13
Electrodes
Process connection
Terminal head
Dissipator
Seals
Insulator body
Stainless steel 1.4571 (AISI 316Ti)
Stainless steel 1.4571 (AISI 316Ti)
Die-cast aluminum
Die-cast aluminum
FFKM
Ceramic
16
Endress+Hauser
Condumax CLS12/CLS13
9.4.3
EU Declaration of conformity
Process connection
CLS12 and CLS13
Thread G1
NPT 1" thread
10
EU Declaration of conformity
Endress+Hauser
17
Index
Index
A
O
Ambient temperature range . . . . . . . . . . . 14
Occupational safety . . . . . . . . . . . . . . . . . 4
Operational safety . . . . . . . . . . . . . . . . . . 4
C
Cell constant . . . . . . . . . . . . . . . . . . . . 14
Certification body . . . . . . . . . . . . . . . . . . 8
Check
Connection . . . . . . . . . . . . . . . . . . 11
Installation . . . . . . . . . . . . . . . . . . . 9
Connection
Check . . . . . . . . . . . . . . . . . . . . . 11
Ensuring the degree of protection . . . . . 11
Connection conditions . . . . . . . . . . . . . . . 10
D
Declaration of conformity . . . . . . . . . . . 8, 17
Degree of protection
Ensuring . . . . . . . . . . . . . . . . . . . . 11
Technical data . . . . . . . . . . . . . . . . 14
Designated use . . . . . . . . . . . . . . . . . . . . 4
Disposal . . . . . . . . . . . . . . . . . . . . . . . 13
E
Electrical connection . . . . . . . . . . . . . . . . 9
Environment . . . . . . . . . . . . . . . . . . . . 14
EU Declaration of conformity . . . . . . . . . . . 17
H
Hazardous area approvals . . . . . . . . . . . . . 8
Hazardous areas . . . . . . . . . . . . . . . . . . . 5
I
Incoming acceptance . . . . . . . . . . . . . . . . 7
Installation
Check . . . . . . . . . . . . . . . . . . . . . . 9
Sensor . . . . . . . . . . . . . . . . . . . . . . 9
M
Manufacturer inspection certificate . . . . . . .
Materials . . . . . . . . . . . . . . . . . . . . . .
Measured values . . . . . . . . . . . . . . . . . .
Measuring ranges . . . . . . . . . . . . . . . . .
. 8
16
14
14
N
Nameplate . . . . . . . . . . . . . . . . . . . . . . 7
P
Pressure-temperature ratings . . . . . . . . . . 15
Process . . . . . . . . . . . . . . . . . . . . . . . . 15
Process connection . . . . . . . . . . . . . . . . . 17
Process pressure . . . . . . . . . . . . . . . . . . 15
Process temperature . . . . . . . . . . . . . . . . 15
Product identification . . . . . . . . . . . . . . . . 7
Product safety . . . . . . . . . . . . . . . . . . . . 5
R
Repair . . . . . . . . . . . . . . . . . . . . . . . . 13
Return . . . . . . . . . . . . . . . . . . . . . . . . 13
S
Safety
Electrical equipment in hazardous areas . . 5
Occupational safety . . . . . . . . . . . . . . 4
Operation . . . . . . . . . . . . . . . . . . . . 4
Product . . . . . . . . . . . . . . . . . . . . . 5
Safety instructions . . . . . . . . . . . . . . . . . . 4
Scope of delivery . . . . . . . . . . . . . . . . . . . 8
Sensor
Cleaning . . . . . . . . . . . . . . . . . . . . 12
Connecting . . . . . . . . . . . . . . . . . . 11
Mounting . . . . . . . . . . . . . . . . . . . . 9
State of the art . . . . . . . . . . . . . . . . . . . . 5
Storage temperature . . . . . . . . . . . . . . . . 14
Symbols . . . . . . . . . . . . . . . . . . . . . . . . 3
T
Technical data
Environment . . . . . . . . . . . . . . . . .
Input . . . . . . . . . . . . . . . . . . . . . .
Mechanical construction . . . . . . . . . .
Process . . . . . . . . . . . . . . . . . . . . .
Temperature compensation . . . . . . . . . . . .
Temperature/pressure ratings . . . . . . . . . .
14
14
16
15
14
15
U
Use . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
W
Warnings . . . . . . . . . . . . . . . . . . . . . . . 3
18
Endress+Hauser
Index
Weight . . . . . . . . . . . . . . . . . . . . . . . . 16
Endress+Hauser
19
www.addresses.endress.com