Emerson Rosemount FPD for Gas Chromatographs Hardware Reference Manual
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
7R00370-H01
Revision B
October 2017
FPD for Gas Chromatographs
Hardware Reference Manual
Introduction
NOTICE
ROSEMOUNT (“SELLER”) SHALL NOT BE LIABLE FOR TECHNICAL OR EDITORIAL ERRORS IN THIS MANUAL OR
OMISSIONS FROM THIS MANUAL. SELLER MAKES NO WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, WITH RESPECT TO THIS MANUAL
AND, IN NO EVENT, SHALL SELLER BE LIABLE FOR ANY SPECIAL OR CONSEQUENTIAL DAMAGES INCLUDING, BUT NOT
LIMITED TO, LOSS OF PRODUCTION, LOSS OF PROFITS, ETC.
PRODUCT NAMES USED HEREIN ARE FOR MANUFACTURER OR SUPPLIER IDENTIFICATION ONLY AND MAY BE
TRADEMARKS/REGISTERED TRADEMARKS OF THESE COMPANIES.
THE CONTENTS OF THIS PUBLICATION ARE PRESENTED FOR INFORMATIONAL PURPOSES ONLY, AND WHILE EVERY
EFFORT HAS BEEN MADE TO ENSURE THEIR ACCURACY, THEY ARE NOT TO BE CONSTRUED AS WARRANTIES OR
GUARANTEES, EXPRESSED OR IMPLIED, REGARDING THE PRODUCTS OR SERVICES DESCRIBED HEREIN OR THEIR USE
OR APPLICABILITY. WE RESERVE THE RIGHT TO MODIFY OR IMPROVE THE DESIGNS OR SPECIFICATIONS OF SUCH
PRODUCTS AT ANY TIME.
SELLER DOES NOT ASSUME RESPONSIBILITY FOR THE SELECTION, USE, OR MAINTENANCE OF ANY PRODUCT.
RESPONSIBILITY FOR PROPER SELECTION, USE, AND MAINTENANCE OF ANY SELLER PRODUCT REMAINS SOLELY WITH
THE PURCHASER AND END-USER.
THE EMERSON LOGO IS A TRADEMARK AND SERVICE MARK OF EMERSON ELECTRIC CO. ROSEMOUNT IS A TRADEMARK
OF ONE OF THE EMERSON FAMILY OF COMPANIES.
©2017
ROSEMOUNT
HOUSTON, TX
USA
All rights reserved. No part of this work may be reproduced or copied in any form or by any means–graphic, electronic,
or mechanical–without first receiving the written permission of Rosemount Houston, Texas, U.S.A.
WARRANTY
LIMITED WARRANTY: Subject to the limitations contained in Section 2 herein and except as otherwise expressly
provided herein, Rosemount (“Seller”) warrants that the firmware will execute the programming instructions
provided by Seller, and that the Goods manufactured or Services provided by Seller will be free from defects in
materials or workmanship under normal use and care until the expiration of the applicable warranty period. Goods are
warranted for twelve (12) months from the date of initial installation or eighteen (18) months from the date of
shipment by Seller, whichever period expires first. Consumables and Services are warranted for a period of 90 days
from the date of shipment or completion of the Services. Products purchased by Seller from a third party for resale to
Buyer (“Resale Products”) shall carry only the warranty extended by the original manufacturer. Buyer agrees that Seller
has no liability for Resale Products beyond making a reasonable commercial effort to arrange for procurement and
shipping of the Resale Products. If Buyer discovers any warranty defects and notifies Seller thereof in writing during
the applicable warranty period, Seller shall, at its option, promptly correct any errors that are found by Seller in the
firmware or Services, or repair or replace F.O.B. point of manufacture that portion of the Goods or firmware found by
Seller to be defective, or refund the purchase price of the defective portion of the Goods/Services. All replacements or
repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuitable
environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any
other cause not the fault of Seller are not covered by this limited warranty, and shall be at Buyer's expense. Seller shall
not be obligated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in
writing in advance by an authorized Seller representative. All costs of dismantling, re-installation and freight and the
time and expenses of Seller's personnel for site travel and diagnosis under this warranty clause shall be borne by Buyer
unless accepted in writing by Seller. Goods repaired and parts replaced during the warranty period shall be in warranty
for the remainder of the original warranty period or ninety (90) days, whichever is longer. This limited warranty is the
only warranty made by Seller and can be amended only in a writing signed by an authorized representative of Seller.
Except as otherwise expressly provided in the Agreement, THERE ARE NO REPRESENTATIONS OR WARRANTIES OF ANY
KIND, EXPRESSED OR IMPLIED, AS TO MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE, OR ANY OTHER
MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES. It is understood that corrosion or erosion of materials is
not covered by our guarantee.
LIMITATION OF REMEDY AND LIABILITY: SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DELAY IN
PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WARRANTY HEREUNDER SHALL BE LIMITED TO
REPAIR, CORRECTION, REPLACEMENT, OR REFUND OF PURCHASE PRICE UNDER THE LIMITED WARRANTY CLAUSE IN
SECTION 1 HEREIN. IN NO EVENT, REGARDLESS OF THE FORM OF THE CLAIM OR CAUSE OF ACTION (WHETHER BASED
IN CONTRACT, INFRINGEMENT, NEGLIGENCE, STRICT LIABILITY, OTHER TORT OR OTHERWISE), SHALL SELLER'S
LIABILITY TO BUYER AND/OR ITS CUSTOMERS EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS
MANUFACTURED OR SERVICES PROVIDED BY SELLER GIVING RISE TO THE CLAIM OR CAUSE OF ACTION. BUYER
AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY TO BUYER AND/OR ITS CUSTOMERS EXTEND TO INCLUDE
INCIDENTAL, CONSEQUENTIAL, OR PUNITIVE DAMAGES. THE TERM “CONSEQUENTIAL DAMAGES” SHALL INCLUDE,
BUT NOT BE LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVENUE, AND COST OF CAPITAL.
Introduction
Table of Contents
Table of Contents
1
Introduction .......................................................................... 1
1.1
1.2
1.3
2
Setup................................................................................... 19
2.1
2.2
2.3
2.4
3
3.2
3.3
Model 500 FPD drawings ............................................................................................. 29
Model 700 FPD drawings ............................................................................................. 30
Enclosure threaded entry details ................................................................................. 51
Appendix B: Manufacturer’s manuals ..................................... 57
B.1
B.2
C
Operation.................................................................................................................... 23
3.1.1 Igniting the flame manually ............................................................................. 24
Maintenance ............................................................................................................... 25
Troubleshooting.......................................................................................................... 26
Appendix A: Drawings .......................................................... 29
A.1
A.2
A.3
B
Gas connections .......................................................................................................... 19
EnvironmentalConsiderations ..................................................................................... 19
Utility gases ................................................................................................................. 19
Venting ....................................................................................................................... 20
Operation and maintenance.................................................. 23
3.1
A
Theory of operation ........................................................................................................2
Equipment description ...................................................................................................4
1.2.1 Model 500 FPD ...................................................................................................6
1.2.2 Model 700 FPD ...................................................................................................9
1.2.3 Model 700 FPD front entry ............................................................................... 12
1.2.4 Model 700XA FPD ............................................................................................ 13
1.2.5 Model 700XA FPD front entry .......................................................................... 14
1.2.6 Model 1500XA FPD .......................................................................................... 15
Glossary ...................................................................................................................... 16
FlamePhotometric Detector Operation Manual ........................................................... 58
GUB FPD 118500-3411 GUB ........................................................................................ 67
Appendix C: Spare Parts List ................................................ 107
i
Introduction
1
Introduction
The flame photometric detector (FPD) that you have received is factory-engineered to be used in
conjunction with all Rosemount gas chromatographs. The FPD can be used as a solitary detector to
measure low levels of sulfur compounds in natural gas or as a secondary detector in conjunction
with a thermal conductivity detector (TCD) that allows the GC to analyze the full range of
components present in a natural gas sample, including sulfur compounds.
Figure 1-1: Flame Photometric Detector (FPD)
A. Flame photometric detector (FPD)
B.
Photomultiplier tube
C. Flame cell
D. Electrometer board
An FPD (A) typically consists of the following major components:
The flame cell (C) - Located in the lower enclosure, the flame cell has connections for fuel gas,
hydrocarbon-free air,sample injection (process gas plus nitrogen carrier), and an exhaust pipe. It is
fitted with an RTD to monitor the temperature when running, and an ignitor to light the fuel gas.
The photomultiplier tube (B) - Located in the lower enclosure, the photomultiplier tube contains the
sensors that measure the light that is emitted from the flame cell during operation. It has one signal
lead and one high voltage wire that take the signal from the detector to the electrometer board
and provide the power for ignition. The leads are co-axial typecables.
1
Introduction
The electrometer board (D) - Located in the upper enclosure, the electrometer board amplifies and
processes the signal data from the detector, and sends it to the CPU board on the GC. It also
provides the ignition circuit, controls the re-light function, and generates the flame out alarm.
1.1
Theory of operation
NOTICE
See also Section 1.3 of this manual, for definitions of some of the terminology used in the following
explanations.
The detection system in the FPD uses the reactions of sulfur components in a hydrogen or air flame
as a source for analytical detection. The source of the FPD's signal is derived from the light
produced by an excited molecule created in the flame's combustion, which is a photochemical
process called chemiluminescence.
A thermocouple is fitted to the flame cell to ensure that the flame is present. If the flame is not
detected, the electrometer shuts off the hydrogen to the flame cell. It then supplies a voltage to
the igniter, waits five seconds and opens the hydrogen shut off valve. The electrometer will make
ten ignition attempts if necessary. If it is not successful, then the hydrogen is shut off, an alarm is
triggered on the GC and the unit awaits attention from the operator.
NOTICE
To ignite the flame manually, see Section 3.1.1.
The signal is sent from the PMT to the electrometer to be amplified. The electrometer also provides
the PMT with the high voltage it requires to operate the auto re-light circuits.1.3
2
Introduction
Figure 1-2: Elution of Components
1.
Carrier gas only at the detector
2.
First component begins to elute from the columns and is sensed by the detector.
3.
Peak concentration of first component.
4.
The second component begins to elute from the columns and is sensed by the detector.
5.
Peak concentration of the second component.
3
Introduction
The signal is then sent to the preamplifier board for further amplification. In addition, the
preamplifier converts each voltage signal to a value that is proportional to the concentration of the
component detected in the gas sample. The preamplifier provides four different gain channels as well
as compensation for baseline drift. The signals are sent to the GC for computation or for viewing on
a PC monitor or local operator interface (LOI).
While the GC is in Idle mode, prior to injecting a sample, the detector is exposed to pure carrier gas.
In this condition, the output from the detector is electrically nulled. The detector output is set to 1
mV DC. This is measured on the red and black terminals on the preamplifier board, and adjusted
using the potentiometer (R38) on the electrometer PCB.
1.2
Equipment description
FPDs are available in the following configurations:
•
Model 500 FPD
•
Model 700 FPD
•
Model 700 FPD Front Entry
•
700XA FPD
•
700XA FPD Front Entry
•
1500XA FPD
NOTICE
The Front Entry configurations include an additional frame to allow all the FPD enclosures to be
mounted on the front of the unit. This allows the unit to be located close to a wall because no rear
access is required for installation or maintenance.
All configurations are ATEX-certified. The differences between the configurations are detailed in
later sections of this chapter.
4
Introduction
The FPD used with the Model 500, 700, and 700XA gas chromatographs has the following
hazardous area certification markings:
5
Introduction
1.2.1
Model 500 FPD
Figure 1-3: Model 500 FPD
The Model 500 FPD module consists of three Exd GUB enclosures mounted on a frame, plus an Exd
solenoid that acts as a hydrogen shut-off valve. The enclosures contain the following:
6
•
Electrometer assembly in GUB 5 enclosure
•
Flame cell and photometric detector tube in GUB 5 enclosure
•
Transformer (either 230/110 Vac or 110/110 Vac) in GUB 4 enclosure
•
Hydrogen shut-off valve
Introduction
Figure 1-4: Model 500 FPD Enclosures
A. Electrometer assembly
B.
Flame cell and FP tube
C. H2 shut off valve
D. Transformer
Place the FPD module as close as possible to its partner GC to minimize the length of sample tubing
between them, and therefore to keep the cycle time as short as possible.
The tubing required to operate the FPD flame cell is 1/16 in. OD 0.010 in. ID. All tubing enters the
GUB enclosure containing the flame cell via a specially designed tubing gland. All internal fittings
are Swagelok double ferrule type compression fittings.
7
Introduction
Figure 1-5: Specialized Tubing Gland
8
Introduction
1.2.2
Model 700 FPD
Figure 1-6: Model 700 FPD
The Model 700 FPD module consists of four Exd GUB enclosures mounted on a frame, plus an Exd
solenoid valve that acts as a hydrogen shut-off valve. The Model 700 FPD requires an additional
enclosure to house temperature control equipment that on a Module 500 GC is available internally.
9
Introduction
The enclosures contain the following:
•
Electrometer assembly in GUB 5 enclosure
•
Flame cell and photometric detector tube in GUB 5 enclosure
•
PID Temperature controller and relay
•
Transformer (either 230/110 Vac or 110/110 Vac) in GUB 4 enclosure
•
Hydrogen shut-off valve
Figure 1-7: PID Temperature Controller and Relay
Place the FPD module as close as possible to its partner GC to minimize the length of sample tubing
between them, and therefore to keep the cycle time as short as possible.
The tubing required to operate the FPD flame cell is 1/16 in. OD 0.010 in. ID. All tubing enters the
GUB enclosure containing the flame cell via a specially designed tubing gland. All internal fittings are
Swagelok double ferrule type compression fittings.
10
Introduction
Figure 1-8: Specialized Tubing Gland
11
Introduction
1.2.3
Model 700 FPD front entry
Figure 1-9: Model 700 FPD Front Entry
The Model 700 FPD front entry is comprised of the same components as the standard Model 700
FPD with an additional frame added to allow all the enclosures to be mounted on the front of the
unit. This allows the unit to be located close to a wall because no rear access is required for
installation ormaintenance.
12
Introduction
1.2.4
Model 700XA FPD
Figure 1-10: Model 700XA FPD
The Model 700XA FPD consists of four explosion-proof enclosures mounted on a frame plus an
explosion-proof solenoid valve that acts as a hydrogen shut-off valve.
13
Introduction
The enclosures contain the following components:
•
Electrometer assembly
•
Flame cell and photometric detector tube
•
Transformer, either a 230/110 Vac or a 110/110 Vac
•
PID temperature controller and relay
•
Hydrogen shut-off valve
Place the FPD as close as possible to its partner GC to minimize the length of sample tubing
between them, and keep the cycle time as short as possible.
The tubing size required to operate the FPD flame cell is 1/16 in. OD 0.010 in. ID. All tubing enters
the flame cell’s enclosure through a specially designed tubing gland. All internal fittings are
Swagelok double ferrule compression fittings.
Figure 1-11: Specialized Tubing Gland
1.2.5
Model 700XA FPD front entry
The Model 700XA FPD front entry is comprised of the same components as the standard Model
700XA FPD with an additional frame added to allow all the enclosures to be mounted on the front
of the unit. This allows the unit to be located close to a wall, because no rear access is required for
installation or maintenance.
14
Introduction
1.2.6
Model 1500XA FPD
Figure 1-12: Model 1500XA FPD
The Model 1500XA FPD consists of four explosion-proof enclosures mounted on a frame plus an
explosion-proof solenoid valve that acts as a hydrogen shut-off valve. The enclosures contain the
following components:
•
Electrometer assembly
•
Flame cell and photometric detector tube
•
Transformer, either a 230/110 Vac or a 110/110 Vac
•
PID temperature controller and relay
•
Hydrogen shut-off valve
15
Introduction
Place the FPD as close as possible to its partner GC in order to minimize the length of sample tubing
between them, and therefore to keep the cycle time as short as possible.
The tubing size required to operate the FPD flame cell is 1/16 in. OD 0.010 in. ID. All tubing enters
the flame cell’s enclosure through a specially designed tubing gland. All internal fittings are
Swagelok double ferrule compression fittings.
Figure 1-13: Specialized Tubing Gland
1.3
Glossary
Auto Zero
Automatic zeroing of the preamplifier. May be entered into the controller to take place at any time
during the analysis when either the component is not eluting or the baseline is steady (not
normally used).
Chromatogram
A permanent record of the detector output. A chromatograph is obtained from a PC interfaced with
the detector output through the GC controller. A typical chromatogram displays all component peaks
and gain changes. It may be viewed in color as it is processed on a PC VGA display. Tick marks
recorded on the chromatogram by the GC Controller indicate where timed events takeplace.
Component
Any one of several different gases that may appear in a sample mixture. For example, sample gas
usually contains the following components: ethyl mercaptan, t-butyl mercaptan, methyl ethyl
sulphide, diethyl sulphide, hydrogen sulphide, and carbonyl sulphide.
16
Introduction
Response Factor
Correction factor for each component as determined by the calibration. It is defined by the
equation:
where
ARFn = Area response factor for component n in area per mole percent (%)
HRFn = Height response factor for component n
Arean = Area associated with component n in calibration gas
Htn = Height associated with component n in mole % in calibration gas
Caln = Amount of component n in mole % in calibration gas
Retention time
The time in seconds that elapses between the start of analysis ( 0 seconds) and the sensing of the
maximum concentration of each component by the analyser detector.
17
Introduction
18
Setup
2
Setup
2.1
Gas connections
2.2
®
Use Silcosteel or equivalent tubing for all calibration gas and process gas connections on all FPDs
that are used to measure low range sulfur components. If you use Grade 316 or other stainless steel
piping, the sulfur components will adhere to the internal surface of the pipe, and will continue to
do so until the entire internal surface of the tubing is coated or conditioned, which will result in
lower than expected levels of sulfur components reaching the detector for measurement.
Conditioning may take one week or longer, depending on the levels of sulfur components and the
length of the tubing.
Environmental Considerations
FPDs are sensitive to changes in temperature and pressure; therefore, place them in shelters that
have stable temperature and pressure. Do not use positive pressurization for shelters.
2.3
Utility gases
FPDs require the following utility gases:
•
Hydrogen - 99.995% purity
•
Hydrocarbon-free air
•
Nitrogen - 99.995% purity (carrier gas)
•
Helium - 99.995% purity (optional second carriergas)
19
Setup
®
Make all utility and process gas connections with Swagelok 1/8-inch double ferrule compression
fittings. Metric conversion kits are available; contact your Rosemount sales representative for more
information.
These are typical values supplied for information only. Actual values are application specific.
Figure 2-1: Typical Pressure and Flow Rate Information
2.4
Venting
All Rosemount FPD modules have a vent from the flame cell that exits the GUB enclosure via a
proprietary Exd breather/drain/ flame arrestor assembly. The exhaust from the flame cell emits
water vapor as a result of burning hydrogen as fuel. This vapor condenses in the exhaust tubing
outside the GUB enclosure, and can be seen as drips of water.
Vent the FPD exhaust to atmosphere. Do not subject the vent to any back pressure because this will
have a detrimental effect on the detector, and may cause the flame to extinguish.
20
Setup
WARNING
Hydrogen-air mixtures can ignite with very low energy input. For reference, an invisible spark or a
static spark from a person can cause ignition. Although the auto- ignition temperature of hydrogen
is higher than those for most hydrocarbons, hydrogen's lower ignition energy makes the ignition of
hydrogen–air mixtures more likely.
Use a container with the FPD module to collect the condensed water from the FPD vent. Do not
pipe the vent away unless you can guarantee a continuous downward slope on the pipe and no
back pressure or obstruction by water.
21
Setup
22
Operation and maintenance
3
Operation and maintenance
3.1
Operation
The FPD operates as a separate detector. It is controlled by and reports to the GC. The flow rates for
the utility gases and the carrier gas are factory set, and are specific to each FPD. These should only
be adjusted by fully trained and authorized personnel.
The FPD is identified as Detector #1 on the Detector screen, which is viewable with MON2000,
MON2020, and the LOI. When used in conjunction with a TCD, the FPD is Detector #1, and the TCD
is Detector #2.
NOTICE
The electrometer switch (A), which has three positions—up for Reset, centered for Normal, and
down for Override—should not be left in Override.
A. Electrometer switch
3-1
23
Operation and maintenance
3.1.1
24
Igniting the flame manually
1.
Connect air to the inlet and slowly bring the inlet pressure to 60 psig.
2.
Connect hydrogen to the inlet and slowly bring theinlet pressure to 60 psig.
3.
Remove tubing from flame cell exhaust and use a digitalflow meter to adjust the air control
valve until a reading of 160 cc/ min is obtained.
4.
Turn off the air supply.
5.
Set the auto relight switch (S1) on the electrometer PCB to the OVERRIDE position.
6.
Use the digital flow meter to adjust the hydrogen controlvalve until a reading of 100
cc/min is obtained.
7.
Turn on the air supply.
8.
Set the auto relight switch (S1) on the electrometer PCB to the RUN position. The auto
relight sequence starts as follows:
a.
The LED on the electrometer comes on after 10 seconds, and the glow plug fitted to the
side of the flame cell is now supplied a voltage.
b.
After another 5 seconds, the hydrogen shut off valve operates.
c.
The gas mixture is ignited.
d.
If the flame does not light in 5 seconds, the electrometer de-energizes the hydrogen
shut off valve to stop the flow into the flame cell.
e.
The flame cell is then purged with air and nitrogen carrier.
f.
The process starts again—up to 10times—until the flame stays lit.
Operation and maintenance
g.
If the flame does not stay lit, the LED flashes. If the alarm output is linked to the 2350A
controllerdiscreet input, there will be an alarm present on the controller.
h.
Set the auto relight switch (S1) on the electrometer PCB to the RESET position and then back
to the RUN position. The re-light sequence isrestarted.
If the unit still fails to light after resetting the electrometer, recheck the air and hydrogen flows.
3.2
Maintenance
The FPD is a complex piece of equipment and needs to be regularly maintained, preferably as part
of an annual planned maintenance process.
The following important maintenance procedures should be conducted on an annual basis:
Replace the flame cell and photometric tube O-rings, except for the Kalrez O-ring, which
should be replaced every 24 months.
•
CAUTION
Be certain that the flame cell has cooled down before touching it, because it often reaches a
temperature of 170 °C (338 °F).
•
Lubricate the stem of the hydrogen shut-offvalve.
For both of these operations, the GC should be shut down, and the appropriate permits and
permissions gained before commencing.
Only trained and authorized personnel should carry out maintenance..
NOTICE
The flame out logic should always be tested to ensure it works at the end of any maintenance.
Failure to maintain the FPD may cause a loss of functionality and can result in permanent damage to
the equipment.
25
Operation and maintenance
3.3
Troubleshooting
Only competent trained personnel should troubleshoot FPDs.
The following list of faults is not definitive. It only details the most common faults.
Fault symptom(s)
•
Check high voltage is present on coax.
•
Approx. -600 Vdc
•
If voltage now present on board, check coax cable.
•
Check BNC coax connectors are tight.
If voltage is present, check
signal coax.
•
If there is no voltage, or the signal cable is OK, replace electrometer.
Upsets are being seen, but there
are no peaks when gas is
injected.
•
Check the 12 V GND wiring to the electrometer board. The two GND terminals on
connector #2 are not linked on board. If there are three black wires, ensure that pins 1
and 4 are connected to the power supply. The other wire is for the flame cell GND.
•
Check the tubing going into the bottom of the flame cell. Loosen fitting and pull
tubing downwards while watchingCGM.
•
If peaks appear, then the tubing needs to be cut.
•
Check to see if there is flow from the metering valve next to the heater block.
•
Check the sample is getting to the flame cell.
•
Try replacing the columns one at a time.
•
Check you are getting carrier through port 1 with valve 2 on and through port 5 with
valve 2 off. If not check the vents on the Alcon valve for back pressure.
When monitoring the baseline
in MON2020, there are no
upsets present when the auto
re- light circuit fires.
If no voltage, remove coax
connector.
26
Possible solution(s)
Operation and maintenance
Fault symptom(s)
Air and H2 flows are set
correctly, and the unit fails to
stay lit.
Possible solution(s)
•
Using a digital thermometer connected to the thermocouple wires
coming from the bottom of the flame-cell, check that the temperature is
160 °C (320 °F).
•
Check flame out thermocouple wires.
•
Ensure no insulation is trapped under screw on terminalstrip.
•
Try pulling the sample tube out when it is attempting to light in case the
tube is affecting the fuel mixture.
•
Replace the flame cell and try again.
•
Ensure that the signal wires are connected to the correct place;
remember that the white signal wire should be connected tothe TC+ of
the CON5.
Unit give good size sample
peaks; then after a while, the
peaks are not present, but the
relight still gives good peaks.
There might be soot on the sample tube going to the flame-cell. Pull tube down
slightly while watching the CGM to see if that cures the fault.
Flame cell temperature cannot
be controlled.
Check the flame cell thermistor.
The resistance is approximately 100 KΩ at ambient. Resistance goes down as
temperature goes up.
•
Check that the thermistor has not been pushed right through the flame
cell.
•
In later models, the flame cell will be blanked at end of holes to ensure
that this cannot happen.
•
Check there is enough heat-sink compound fitted around sensors.
•
Check the BNC connectors for the signal in and the high voltage. Ensure
that they are tight.
•
Cut off the flame and check the response from the detector on a live
CGM.
•
Try changing the filter.
Restrictor metering valve seems
to be restricting the output flow
completely.
•
Apply Snoop® to the two fittings at the bottom of the metering valve.
•
Change the metering valve.
Peaks are very small or appear
to be back to front.
•
Check nitrogen flow into union at flame cell.
•
This should be no less than 15cc/min.
Flame cell temperature is
erratic.
Unable to balance the bridge.
27
Operation and maintenance
28
Fault symptom(s)
Possible solution(s)
Noisy baseline and/or very big
dips on the baseline.
Check the air supply, which should be no lower than 500psi in the cylinder.
Appendix A: Drawings
A
Appendix A: Drawings
A.1
Model 500 FPD drawings
DUK 7233/013/1
General Arrangement: Model 500 FPD Module
DUK 7233/002/1
General Arrangement: Model 500 FPD Analyzer
DUK 7233/039/1
General Arrangement: Model 500 FPD Dual Analyzer
DUK 7233/028/1
Power Wiring Diagram: 500 FPD C/W Aux Stream Switching
DUK 7233/029/1
Wiring Diagram: FPD Relight Failure Alarm
DUK 7233/030/1
Power Wiring Diagram: Model 500 FPD
DUK 7233/033/1
Interconnection Diagram: 500 FPD / 2350A Controller
DUK 7233/034/1
Wiring Diagram: 500 FPD / 2350A Controller
DUK 7233/048/1
Wiring Diagram: 500 FPD / 2350A – 6 x 6 Port Valves
DUK 7233/056/1
Power Distribution : 500 FPD Dual Analyser C/W 2 x Trace Heat
DUK 7233/062/1
Wiring Diagram : 500 FPD / 2350A – 6 x 6 Port – Aux Stream Switching
NOTICE
Flow Diagrams: Refer to the sales order documentation that was shipped with the GC.
29
Appendix A: Drawings
A.2
Model 700 FPD drawings
DUK 7204/074/1
General Arrangement: 700 FPD Analyzer
DUK 7204/100/1
General Arrangement: 700 FPD Module
DUK 7204/156/1
General Arrangement: 700 FPD Module Front Entry
DUK 7204/102/1
Internal Cable Wiring: 700 FPD Analyzer
DUK 7204/103/1
Power Wiring Diagram: 700 FPD Analyzer
NOTICE
Flow Diagrams: Refer to the sales order documentation that was shipped with the GC.
30
Appendix A: Drawings
31
Appendix A: Drawings
32
Appendix A: Drawings
33
Appendix A: Drawings
34
Appendix A: Drawings
35
Appendix A: Drawings
36
Appendix A: Drawings
37
Appendix A: Drawings
38
Appendix A: Drawings
39
Appendix A: Drawings
40
Appendix A: Drawings
41
Appendix A: Drawings
42
Appendix A: Drawings
43
Appendix A: Drawings
44
Appendix A: Drawings
45
Appendix A: Drawings
46
Appendix A: Drawings
47
Appendix A: Drawings
48
Appendix A: Drawings
49
Appendix A: Drawings
50
Appendix A: Drawings
A.3
Enclosure threaded entry details
DUK 7233/060/3
FPD Module Bottom Housing Assembly
DUK 7233/061/3
FPD Module Top Housing Assembly
DUK 7204/101/3
Model 700 FPD Module Temperature/Controller Enclosure
DUK 7233/007/3
FPD Module Transformer Housing Assembly
51
Appendix A: Drawings
52
Appendix A: Drawings
53
Appendix A: Drawings
54
Appendix A: Drawings
55
Appendix A: Drawings
56
Appendix B: Manufacturer’s manuals
B
Appendix B: Manufacturer’s manuals
•
Flame Photometric Detector Operation Manual
•
PID Controller Manual
57
Appendix B: Manufacturer’s manuals
B.1
FlamePhotometric Detector Operation Manual
23332-K026
Revision B April 25th, 2008
Information disclosed herein may not be reproduced in any form without the express
permission of GC EXPRESS.
58
Appendix B: Manufacturer’s manuals
IMPORTANT
In order to obtain optimum performance from this detector, it is necessary to meet and
maintain the following conditions:
A. The following minimum purity standard for gases and liquids shall be maintained:
Helium - 99.999% (ultra high purity) or Nitrogen - 99.999% (ultra high purity) Hydrogen 99.999% (ultra high purity)
Air - 0.1 PPM total Hydrocarbons (ultra zero grade)
B. Stainless steel diaphragm regulators must be used.
C. All gas lines from source to instrument must be clean.
B.1.1
General Description
Introduction
The Flame Photometric Detector, FPD, is a very sensitive and selective detector for the
analysis of sulfur or organophosphorus containing compounds. The detector is very stable
and easy to use. As the analyte is burned in a hydrogen and air flame, a characteristic
wavelength of light is emitted at 394 nm for sulfur and 526 nm for phosphorus. A filter
specific to the appropriate wavelength may be installed to enhance the selectivity to the
sulfur or phosphorus emission. The emitted light is amplified by the photomultiplier tube
(PTM) and processed by the signal processor. 'The response to phosphorus is linear and
quadratic to sulfur.
The detector may be operated in either the sulfur mode or phosphorus mode by switching
the filter and adjusting the air to hydrogen ratio to optimize response. A shielded flame
design of the detector enhances sensitivity by lowering the noise created by the light
emitted by the flame.
The detector uses a stainless steel jet, quartz windows, and silicone 0-rings in an all
aluminum body.
59
Appendix B: Manufacturer’s manuals
Specifications
•
Maximum operating temperature: 250 °C
•
Shielded stainless steel jet
•
Sensitivity: 2 x 10·12 g/sec for sulfur
•
Sensitivity: 1 x 10·12 g/sec for phosphorus
•
Linear range: l 04 for phosphorus
•
Linear range: 103 with optional square root function for sulfur
•
Leak tight design to allow measurement of all flows from detector exhaust
•
Igniter voltage: 1.5V AC at 4 amps
•
PMT voltage variable from approximately 650V
Installation of the FPD Optical filter
In order to have the specificity for sulfur or phosphorus detection the appropriate optical
filter must be in place. The phosphorus filter is a filter of 526 nanometers and the sulfur is a
filter of 396 nanometers.
Before changing the filter, the power cable to the photomultiplier tube, PMT, must be
removed. This will prevent irreparable damage which can be caused by the introduction of
room light to the PMT. The two thumb screws securing the PMT to the detector body are
removed and then the PMT is slid off gently. Some resistance is felt due to the O-ring on
the detector body which provides a light tight seal.
The filter may be removed and replaced with the appropriate filter. The sulfur filter is a
very dark blue color and the phosphorus filter a fluorescent yellow green. One side of the
filter has a mirror finish. There is not a front or back face to the filter.
The PMT is slid back in place and the two thumb screws secured to the detector body.
Reattach the power mid signal cables to the back of the PMT.
60
Appendix B: Manufacturer’s manuals
Figure B-1: Insertion of Capillary Column into the FPD
Operation
CAUTION
When working with the detector, never remove the photomultiplier from the detector with the
dynode voltage applied. Exposure to high light levels will cause photocathode fatigue
(sensitivity loss for an extended period of time) and may cause permanent damage.
Optimal detector temperature
The FPD may operate up to a temperature of 250 °C. Take care to operate the detector
above the final temperature of the column to prevent the condensation of column bleed on
the surface of the optical windows which could result in loss of response.
CAUTION
Do not operate the detector above 250 °C or you may damage the plastic photomultiplier
housing.
Optimizing flows and igniting the flame
The optimization of the detector is achieved by adjusting the ratio of hydrogen to air. The
oxygen content of air should be 0.2 - 0.4 of the hydrogen flow, with the optimum ratio
being 0.3. The air flow should be 1.5 times the hydrogen flow. When optimizing
conditions, the higher the total gas flows; the higher the background noise.
61
Appendix B: Manufacturer’s manuals
Example:
Hydrogen flow 100 mL/min
100 mL/min x 1.5 = 150 mL/min air required
Nitrogen is the most common carrier gas used for packed columns. Helium is used for the
carrier gas for capillary column s with nitrogen for the make-up gas.
Once the flows are set and the detector is at a temperature of at least 125 °C, the flame
may be lit.
Selecting the Linear or Square Root Mode of Operation
The FPD electrometer has two modes of operation designated as" linear" and "square root".
To select the mode of operation, use the sq rt / linear switch.
In the linear mode, the circuit performs as a basic electrometer giving a 10 volt output for
an input current of one microampere. This 10 volt full scale output is available at the 10
volt output. A 1 volt output is also available. The linear mode is used when the detector is
operated in the phosphorus mode of operation with the phosphorus filter installed.
Phosphorus is detected as POH.
62
Appendix B: Manufacturer’s manuals
Sulfur is detected as S2 and the response is proportional to the square of the concentration of
the sulfur containing compound. The square root mode is selected from the switch
marked sq rt / linear. In this mode the electrometer output is modified by a special resistordiode matrix to correct for the non-linear (approximately square law) relationship between
the detector output current and sulfur concentration when the detector is operated in the
sulfur mode. When operating in the mode, the electrometer zero control should be set to
provide a slightly positive output from the module with the detector output at baseline.
Figure B-2: Replacement of O-rings and Windows
63
Appendix B: Manufacturer’s manuals
Maintenance
CAUTION
When working with the detector, never remove the photomultiplier from the detector with the
dynode voltage applied. Exposure to high light levels will cause photocathode fatigue
(sensitivity loss for an extended period of time) and may cause permanent damage to the PMT.
Cleaning the Detector
Column bleed may build up in the FPD housing. This stationary phase coating may be rinsed
out of the detector without disassembly. Follow the procedure listed below:
1. Disconnect the electrical connections from the detector.
2. Tum off the hydrogen and air supply lines to the GC.
3. Cool the detector to ambient.
4. Disconnect the column, hydrogen and air lines from the detector body.
5. Remove the detector from the GC.
6. Cap the hydrogen and air inlets with an 1/8" cap nut.
7. Flush the detector thoroughly with acetone through the column inlet port and exiting
through the exhaust tube.
8. Dry the detector with nitrogen thoroughly.
9. Uncap the gas inlets and reinstall the detector onto the GC.
Replacing the O-rings and Quartz Windows
After using the detector for about twelve months at 250° or more, the O -rings may
become brittle and begin to allow light to leak into the detector resulting in high
background noise and loss of response. The quartz window may need to be replaced as
well. There are a total of five O-rings in the O-ring replacement kit,
PIN 116910-KAL REZ. Four are Kalrez and one is Teflon. The locations of these O- rings are
shown in Figure B-2. These O -rings must be replaced any time a joint sea led by one of
them is separated. ' The cross section view of the detector is shown in Figure B-2.
There are two concentric O-rings between the window housing and flame base. A 1-1/4"
Kalrez ring fits into a groove in the window housing itself and a 15/16" Teflon ring fits
around a bushing between the window and the flame base. A 15/16" Kalrez ring is used
between the window at the inner end of the filter housing and the heat radiator section.
The following procedure should be used to replace the O -rings and quartz windows:
1. Disconnect the power cable from the PMT.
2. Loosen the two thumb screws on the filter housing and remove the PMT.
3. Remove the filter.
4. With a Phillips screwdriver, disconnect the heater-igniter wiring bracket from the
housing assembly.
5. Pull the filter housing from the recess in the heat radiator, exposing first the window and
O-ring (15/16 in. Kalrez).
6. With a hex (Allen) wrench, remove the four screws holding the radiator and window
housing to the flame base.
7. Remove the heat radiator, window housing, and the window
8. Remove the old O-rings.
64
Appendix B: Manufacturer’s manuals
9. Place the 15/16 in. Teflon ring around the metal bushing.
10. Insert the 1-1/4-in. ring into the groove in the window housing.
11. With the bushing and its ring between the window and the flame base and grooved side of
the window housing toward the flame base, align the window housing with the threaded
holes in the flame base.
12. Replace the heat radiator over the window housing with countersunk holes toward the
outside and aligned with the holes in the window housing and flame base.
13. Replace the Allen head screws and tighten.
14. Place the outer window in the recess in the inner end of the filter housing with the 15/16in. Kalrez O-ring between the window and the heat radiator.
15. Replace the filter housing and the wiring bracket.
16. Replace the filter and the PMT.
The 1 -1/4-in. Kalrez end cover O-ring is located between the end cover and the flame base.
Recommended spare parts
The figure below is a cross-section diagram of the FPD with associated part numbers. The parts
listed below are used in the normal maintenance of the detector.
Description
Part number
Igniter plug (includes O-ring seal) – 1.5 volt
116906-K001
Quartz window
23608-0019
O-ring kit
2 ea. # 2-27 Kalrez (1.437 O.D. x 0.70 dia cross section)
1 ea. #2-21 Kalrez (1.062 O.D. x 0.70 dia cross section)
1 ea. #c2118-021 PFTE (1.062 O.D. x 0.70 cross section)
1 ea. #568-010 Kalrez O-ring # S70010 ignitor
1 ea. #586-011 Kalrez O-ring #S70011 flame out TC
116910-KALREZ
65
Appendix B: Manufacturer’s manuals
66
Appendix B: Manufacturer’s manuals
B.2
GUB FPD
118500-3411 GUB
67
Appendix B: Manufacturer’s manuals
B.2.1
Full function uP controlled FPD Rev G
1.
FPD Processor Control Functions
A.
Power on Initialization.
B.
Reset State. The uP monitors the RESET/RUN/OVERRIDE Switch (R/R/0
Switch) (SW1), suspending any automatic operation until the R/R/ 0
Switch is set to the RUN position. If the R/R/0 Switch is set to the
Override position, the uP continues to be Reset but the fuel valve will
be manually activated . The fuel valve will remain activated until the
R/R/0 Switch is manually switched to either the Reset or Run position.
C.
Igniter and Flame on State. When the R/R/ 0 Switch is set to the R UN
position, the uP attempts to ignite the flame. Tl1e ignition sequence
consists of the following steps.
1. Tum the Igniter Drive and LED (D20) on and wait for 5 sec. This
allows the igniter to reach a temperature that will cause ignition.
2. Open the fuel valve and wait for 15 sec.
3. Tum the Igniter Drive and LED off.
4. Check for Flame On by monitoring the thermocouple temperature
sensor input at connector CON5.
5. If no flame is detected, fuel valve will be closed and the uP will delay
for another 30 sec. before any attempt to retry the ignition
sequence.
6. lf a flame is detected, the uP will continue monitoring the
thermocouple temperature sensor input for a flame on indication,
maintaining the fuel valve on and the LED indicator off.
If the uP does not detect a flame within 10 tries of the Ignition Sequence, it will set the
igniter and fuel solenoid off and indicate a error condition by flashing the LED indicator
(D20) at a steady 2Hz. An external error control signal (External Alarm), which can be
used to drive a remote indicator (LED, Buzzer, Etc.), will be activated at connector
CON3.l. The uP will suspend any other operation until the R/ R/0 switch has been
cycled off and back on or the power has been cycled off and back on.
The uP will enter the Ignition Sequence and will attempt ignition:
a. On power up if the R/R/0 Switch is set to RUN.
b. Anytime the R/ R/0 Switch is cycled from RESET or OVER RIDE to RUN.
c. In nominal operation , whenever the flame has been on and has gone out.
If the flame cannot be started within 10 tries of the Ignition Sequence , the uP will not
try to re-ignite until the R/R/0 Switch has been manually cycled off and back on or the
power has been turned off and back on.
Any time the R/R/0 Switch is cycled from RUN to RESET, the uP will stop fuel flow by
turning the fuel solenoid off. No attempt will be made to restart the flame until the R/
R/0 Switch is returned to the RUN position.
Warning: The R/R/0 Switch is a three-position switch, and once switched to the OVERRIDE
position there is no automatic termination of the fuel valve activation. This feature is used
for set.up of the fuel flow only. To deactivate the fuel valve, the R/R0/ Switch must be
manually switched back to the RUN or RESET positions.
Refer to FPO Firmware Flowchart for detailed outline of uP functions.
68
Appendix B: Manufacturer’s manuals
2.
FPD Electrometer Power Supply:
Use caution. AC Voltage (120 Volts AC) is present and DC Voltage in excess of 600
Volts is generated on the PCB when power is applied.
A. Extemal Power, AC Volts: 120 Volts AC routed thru CON4 is switched by
the Solid State Relay U7 (S101DH2). The Gas Valve/Solenoid is controlled
by this switched AC Voltage signal.
B. External Power, DC Volts: 12 Volt DC to low voltage power connector CON2.
CON2, Pins I & 2, power the low current section of the PCB. CON2, Pins 3 & 4,
power high current circuits (HV Regulator, Igniter, fuel solenoid, etc.).
C. On board low voltage:
1.
An on board DC to DC Converter (U6) generates +/- 15 Volts
2.
A LM4040 Voltage Regulator (U5) generates +5 Volts
D. On board high voltage:
On board high voltage converter generates approximately 650 Volt DC (J4)
3.
FPD Linear Mode Test
A. Set the Linear / Sq. Root Switch (SW3) to Linear Mode
B. During the following test steps, monitor U3.6 output line with an
oscilloscope to check for oscillation or other signs of faulty operation.
C.
With Signal In input connector (J3) open, recorder span set to 1 mV. full
scale and the Zero Switch (SW4) set to OFF, adjust R57 for best output null.
D.
With Signal In input connector (J3) open, set the Zero Switch (SW4) ON.
Adjust the manual zero pot (R38, can be located on the PCB or mounted
on the front panel) completely CW and check for an output of +0.055V to
+0.075V. Adjust the zero pot completely CCW and check for a smoothly
changing voltage output to - l. l 5V to -1. 55V. Return the Zero voltage
control to approx. 0 volts output.
E.
Connect a current source to the Signal In input connector (J3). With a
Voltmeter or recorder, monitor the output at the l OV output pin (CON I. 4).
Change recorder span as necessary to check output range and linearity per
following table.
Current source setting (AMPS)
Recorder reading at direct output
-1 x 10-10
1.0 MV +-2%
-1 x 10-9
10.0 MV +-2%
-1 x 10-8
0.100 V +-2%
-1 x 10
1.0 V +-2%
-7
-1 x 10-6
V +-2%
69
Appendix B: Manufacturer’s manuals
4.
FPD Square Root Mode Test
A. Set the Linear / Sq. Root Switch (SW3) to Square Root Mode
B. Set diode oven temperature adjustment pot (R59) near the center of its range
of adjustment. Monitor U3.6 with oscilloscope for oscillation or other signs of
faulty circuit operation.
C. Connect a variable span recorder or DVM (10 megoluns input impedance
minimum) to the lOV output (CONl. 4), and a current source to the input
connector (J3). Set the ZERO SW to ON.
D. Check electrometer and recorder zeros and carefully rese t if necessary. Refer to
Section 3, FPD Linear Mode Test, for zero set procedure.
E.
Set the ZERO SW to ON, the input current to -4.0 X 10·8 amps and adjust the
diode oven temperature by means ofR59 so that when temperature stabilizes
the recorder or DVM reads 31.56 mV as closely as possible.
F.
Reset input current to zero and note recorder/DVM reading. Return input
current to-4.0 X 1·0 8 and trim diode oven temperature if necessary so that the
difference in recorder/DVMreadings for input currents of zero and -4.0 X 10·8
amps is 31.56mV, plus or minus O. l mv.
G. Check response curve per following table. (lf zero reading falls outside
permitted limits, readjust the offset pot (R57) and repeat previous step.
Current source
setting (AMPS)
5.
Direct output (10
VFS) reading (mV)
-0.00 x 10-11
-0.5 ¸ “Zero” < +0.5
R62 = 191K
R62 = 90.9K
-2.00 x 10-11
“Zero reading”
+.167 + .1
+.335 +-.1
-6.00 x 10-11
“Zero reading”
+.470 + -.15
_.945+-1.5
-1.60 x 10-10
“Zero reading”
+1.16 +-.2
+2.34+-.2
-6.40 x 10
“Zero reading”
+3.34 + -.3
+6.71+-.3
-2.50 x 10-9
“Zero reading”
+7.62 +-.6
+15.3 +-.6
-1.00 x 10-8
“Zero reading”
+15.68 +-.9
+31.5+-.9
-4.00 x 10
-8
“Zero reading”
+31.56 +-.1
+63.4 +-.1
-1.60 x 10-7
“Zero reading”
+64.7 +-3.0
+130 +-3.0
-6.40 x 10-7
“Zero reading”
+129 +-5.0
+260 +5.0
-2.56 x 10
“Zero reading”
+319 +-12
+641 +-12
-10
-6
Noise and Drift Test
A. Conduct test with all shields and covers in place and electrometer operating in
the "square root" mode. (SW3 set to square root) Connect lOMV F.S. recorder
to the Direct Output (CONl.4) with the chart speed set to approx. 0.25
cm/min. Disconnect input cable, tum the Zero Switch (SW4) ON and set the
zero control pot (R38) so that trace is near center of plot.
B. Record data for at least 30 minutes in a stable ambient temperature.
C. Acceptance specifications are as follows:
70
1.
Max. Peak-to-peak noise - 2% of full scale.
2.
Occasional unexplained spikes - no more than one per half hour and not
to exceed 5% full-scale peak height.
3.
Max. Drift - 1.5% full scale during half hour nm.
Appendix B: Manufacturer’s manuals
6.
FPD Thermocouple Temperature Setup.
The thermocouple input at CONS Pin1 and Pin2 will be factory adjusted to operate
with Detector temperatures that range from approximately 150 Degrees C to 200
Degrees C.
R
e
t
u
r
n
O
i
r
e
c
t
t
o
P
o
w
e
r
S
u
p
p
l
y
(
G
N
D
)
S
W
1
8
L
E
D
A
6
5
1
D
V
4
1
V
3
S
W
C
O
M
71
Appendix B: Manufacturer’s manuals
72
Appendix B: Manufacturer’s manuals
73
Appendix B: Manufacturer’s manuals
74
Appendix B: Manufacturer’s manuals
75
Appendix B: Manufacturer’s manuals
76
Appendix B: Manufacturer’s manuals
77
Appendix B: Manufacturer’s manuals
78
Appendix B: Manufacturer’s manuals
79
Appendix B: Manufacturer’s manuals
80
Appendix B: Manufacturer’s manuals
81
Appendix B: Manufacturer’s manuals
82
Appendix B: Manufacturer’s manuals
83
Appendix B: Manufacturer’s manuals
84
Appendix B: Manufacturer’s manuals
85
Appendix B: Manufacturer’s manuals
86
Appendix B: Manufacturer’s manuals
87
Appendix B: Manufacturer’s manuals
88
Appendix B: Manufacturer’s manuals
89
Appendix B: Manufacturer’s manuals
90
Appendix B: Manufacturer’s manuals
91
Appendix B: Manufacturer’s manuals
92
Appendix B: Manufacturer’s manuals
93
Appendix B: Manufacturer’s manuals
94
Appendix B: Manufacturer’s manuals
95
Appendix B: Manufacturer’s manuals
96
Appendix B: Manufacturer’s manuals
97
Appendix B: Manufacturer’s manuals
98
Appendix B: Manufacturer’s manuals
99
Appendix B: Manufacturer’s manuals
100
Appendix B: Manufacturer’s manuals
101
Appendix B: Manufacturer’s manuals
102
Appendix B: Manufacturer’s manuals
103
Appendix B: Manufacturer’s manuals
104
Appendix B: Manufacturer’s manuals
105
Appendix B: Manufacturer’s manuals
106
Appendix C: Spare Parts List
C
Appendix C: Spare Parts List
02122 0020
Hydrogen shut-off valve (Alcon)
02122 0023
O-ring kit for Alcon valve
02122 0044
Hydrogen shut-off valve (Asco)
02122 0057
O-ring kit for Asco valve
4-5000-391
Utility gas regulator (H2 or air)
59551 2097
Heater (flame cell and exhaust breather)
116901-GUB
Photometric tube
115000-0008
Power supply for electrometer board
115003-0001
Electrometer board
116910-KALREZ
O-ring kit for flame cell and detector
116906-0001
Ignitor with Kalrez O-ring
23608-0019
Heat filter
23608-0027
Optical filter
NOTICE
For price & delivery information please contact your local Emerson Sales office, or email
gc.csc@emerson.com.
NOTICE
For spare parts for Model 500 or Model 700 Gas Chromatographs, please refer to the appropriate
GC manual.
107
109
FPD for Gas Chromatographs
Manual
7R00370-H01
October 2017
North America Regional
Office Emerson Automation
Solutions 10241 West Little York,
Suite 2200
Houston, TX 77040 USA
+1 866 422 3683 or +1 713 396 8880
+1 713 468 8175
GC.CSC@Emerson.com
Latin America Regional
Office Emerson Automation
Solutions 1300 Concord Terrace,
Suite 400 Sunrise, FL 33323, USA
+1 954 846 5030
+1 952846 5121
GC.CSC@Emerson.com
Europe Regional Office
Emerson Automation Solutions Europe GmbH
Neuhofstrasse 19a P.O. Box 1046
CH 6340 Baar
Switzerland
+1 954 846 5030
+1 952846 5121
GC.CSC@Emerson.com
Asia Pacific Regional Office
Emerson Automation Solutions Asia Pacific Pte
LTD
1 Pandan Crescent
Singapore 128461
+65 6777 8211
+65 6777 0947
GC.CSC@Emerson.com
Analyticexpert.com
Linkedin.com/company/Emerson-Automation-Solutions
Twitter.com/Rosemount_News
Facebook.com/Rosemount
Middle East and Africa Regional Office
Emerson Automation
Solutions Emerson FZE P.O. Box
17033 Jebel Ali Free Zone - South
2
+971 4 8118100
+971 4 88665465
GC.CSC@Emerson.com
Youtube.com/user/RosemountMeasurement
Google.com/+RosemountMeasurement
The Emerson logo is a trademark and service mark of Emerson Electric Co.
Rosemount and Rosemount logotype are trademarks of Emerson.
All other marks are the property of their respective owners.
© 2017 Emerson. All rights reserved.