Ametek Thermox® AMEVision Display User Interface Installation & User Manual Thermo
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Thermox®
AMEVision Display User Interface
Installation & User Manual
Thermox
PN 9000-165-VE , Rev. K
150 Freeport Road
Pittsburgh, PA 15238
Offices
USA - Pittsburgh
150 Freeport Road
Pittsburgh, PA 15238
Ph. 412-828-9040
Fax 412-826-0399
CHINA
AMETEK Shanghai (SPL)
Room 408, Metro Tower
#30 Tian Yao Qiao Road
Shanghai 200030
Ph. 86 21 6426 8111
Fax 86 21 6426 7054
USA - Delaware
455 Corporate Blvd.
Newark, DE 19702
Ph. 302-456-4400
Fax 302-456-4444
AMETEK Beijing
CITIC Building, Room 2305
19, Jianguomenwai Dajie
Beijing 100004
Ph. 86 10 8526 2111
Fax 86 10 8526 2141
USA - Texas
4903 West Sam Houston Parkway North
Suite A-400
Houston, TX 77041
Ph. 713-466-4900
Fax 713-849-1924
FRANCE
AMETEK Precision Instruments France
Rond Point de l’épine des champs
Buroplus Bat D
78990 Elanbourt
Ph. 33 1 30 68 89 20
Fax 33 1 30 68 89 29
CANADA
2876 Sunridge Way N.E.
Calgary, AB T1Y 7H9
Ph. 403-235-8400
Fax 403-248-3550
GERMANY
AMETEK GmbH
Rudolf-Diesel-Strasse 16
D-40670 Meerbusch
Ph. 49 21 59 91 36 0
Fax 49 21 59 91 3680
MIDDLE EAST – Dubai
AMETEK
PO Box 17067
Jebel Ali Free Zone
Dubai, UAE
Ph. 971-4-881 2052
Fax 971-4-881 2053
SINGAPORE
AMETEK Singapore Pte. Ltd.
No 43, Changi South Avenue 2, #04-01
Singapore 486164
Ph. 65-64842388
Fax 65-64816588
© 2015 AMETEK
This manual is a guide for the use of the Thermox AMEVision Display User Interface. Data herein has been verified and validated and is believed adequate for the intended use of this instrument. If the instrument or procedures are used for purposes over and above the capabilities specified herein,
confirmation of their validity and suitability should be obtained; otherwise, AMETEK does not guarantee results and assumes no obligation or liability.
This publication is not a license to operate under, or a recommendation to infringe upon, any process patents.
2 | AMEVision Host Display
Contents
Offices......................................................................................................................2
Safety Notes.............................................................................................................5
Important Notice To Users.......................................................................................6
Environmental Information (WEEE).......................................................................7
General Safety Summary.........................................................................................8
Product Damage Precautions...................................................................................8
Warranty And Claims...............................................................................................9
Conformity.............................................................................................................10
Installation
Introduction............................................................................................................12
Overview................................................................................................................12
Wiring....................................................................................................................13
General Wiring and Conduit Requirements....................................................13
Mains Supply Connections....................................................................................14
EMC Grounding Method.......................................................................................15
Transient and RFI Interference..............................................................................16
AMEVision Mounting & Dimensions Drawing....................................................17
Controller / User Interface
Overview................................................................................................................18
Software Display and Keypad...............................................................................18
Home Screen..........................................................................................................19
Configuration Menu...............................................................................................21
System Passcode....................................................................................................22
System Name Configuration..................................................................................23
System Time..........................................................................................................24
Communication......................................................................................................25
Sensor Configuration.............................................................................................26
Process Pressure.....................................................................................................27
Analog Output.......................................................................................................28
Alarm Configuration..............................................................................................29
Calibration.............................................................................................................32
Manual Calibration..........................................................................................32
Automatic Calibration.....................................................................................32
Remote Calibration.........................................................................................32
Cal Gas Values Menu.............................................................................................33
Start Calibration Menu..........................................................................................34
Calibrate the Analyzer...........................................................................................35
Perform a Manual Calibration.........................................................................35
Perform an Automatic Calibration..................................................................38
Perform a Remote Calibration........................................................................41
User Manual | 3
Flow Calibration..................................................................................................42
Alarms....................................................................................................................44
Diagnostics............................................................................................................45
Sensor Diagnostics.................................................................................................46
Control Diagnostics...............................................................................................47
Inject Gas...............................................................................................................49
Trend Logging.......................................................................................................50
Cell Diagnostics.....................................................................................................52
Combustibles Detector...........................................................................................52
CH4 Detector..........................................................................................................53
Utility.....................................................................................................................54
Detect Sensor.........................................................................................................55
Parameter Management.........................................................................................56
Firmware Upgrade.................................................................................................57
USB Transfer.........................................................................................................58
Reset Age...............................................................................................................61
About.....................................................................................................................62
Alarm and Status Definitions.................................................................................63
Warnings................................................................................................................64
Replacement Parts List..........................................................................................65
Optional Blow Back Configuration.......................................................................66
French Warnings / Français Avertissements
4 | AMEVision Host Display
Safety Notes
WARNINGS, CAUTIONS, and NOTES contained in this manual emphasize critical
instructions as follows:
An operating procedure which, if not strictly observed, may result in personal
injury or environmental contamination.
An operating procedure which, if not strictly observed, may result in damage
to the equipment.
Important information that should not be overlooked.
NOTE
Electrical Safety
Up to 5 kV may be present in the housing. Always shut down power source(s) before performing maintenance or troubleshooting. Only a qualified electrician should make electrical
connections and ground checks.
Any use of the equipment in a manner not specified by the manufacturer may impair the
safety protection originally provided by the equipment.
Grounding
Instrument grounding is mandatory. Performance specifications and safety protection are
void if instrument is operated from an improperly grounded power source.
Verify ground continuity of all equipment before applying power.
User Manual | 5
Important Notice to Users
The AMEVision Display User Interface is suitable for the following locations:
- Class 1, Zone 2, Group IIC
- Class 1, Division 2, Groups A, B, C, D
No operator serviceable components inside any components of the AMEVision. Refer
servicing to qualified personnel.
Power, input, and output (I/O) wiring must be in accordance with Class I, Division 2 wiring
methods [Article 501-4(b) of the National Electric Code, NFPA 70] and in accordance with
the authority having jurisdiction.
WARNING - EXPLOSION HAZARD - SUBSTITUTION OF COMPONENTS
MAY IMPAIR SUITABILITY FOR CLASS I, DIVISION 2.
WARNING - EXPLOSION HAZARD - WHEN IN HAZARDOUS LOCATIONS, TURN OFF POWER BEFORE REPLACING OR WIRING MODULES.
WARNING - EXPLOSION HAZARD - DO NOT DISCONNECT EQUIPMENT UNLESS POWER HAS BEEN SWITCHED OFF OR THE AREA IS
KNOWN TO BE NONHAZARDOUS.
THIS EQUIPMENT IS SUITABLE FOR USE IN CLASS I, DIVISION 2,
GROUPS A, B, C, AND D OR NONHAZARDOUS LOCATIONS ONLY.
WARNING: EXPOSURE TO SOME CHEMICALS MAY DEGRADE THE
SEALING PROPERTIES OF MATERIALS USED IN THE FOLLOWING
DEVICES: POWER SUPPLY MODULE.
WARNING: EXPLOSION HAZARD. WHEN IN HAZARDOUS LOCATIONS,
TURN OFF POWER BEFORE REMOVING OR REPLACING WIRING,
MODULES, BATTERY OR COMPONENTS UNLESS THE AREA IS KNOWN
TO BE FREE OF IGNITABLE CONCENTRATIONS OF FLAMMABLE
GASES OR VAPORS.
WARNING: SUBSTITUTION OF THE FOLLOWING COMPONENTS MAY
IMPAIR SUITABILITY FOR DIVISION 2 POWER SUPPLY.
WARNING: BATTERYMUST BE REPLACED WITH AMETEK PART NUMBER 1000-652-VE.
The maximum ambient temperature for the unit is 55 °C (131 °F)
The AMEVision Display User Interface is a complex piece of equipment that should only
be serviced by a qualified service technician with expertise in instrument technology and
electrical systems. AMETEK recommends that all equipment requiring service be sent back
to the factory. You should only attempt to repair or service this equipment after receiving
training from an AMETEK/P&AI Division training representative. If you decide to service
this equipment be aware that high voltages, high temperatures, and other potentially hazardous conditions may arise.
6 | AMEVision Host Display
Environmental Information (WEEE)
This AMETEK product contains materials that can be reclaimed and recycled. In some cases the product
may contain materials known to be hazardous to the environment or human health. In order to prevent
the release of harmful substances into the environment and to conserve our natural resources, AMETEK
recommends that you arrange to recycle this product when it reached its “end of life”.
Waste Electrical and Electronic Equipment (WEEE) should never be disposed of in a municipal waste
system (residential trash). The Wheelie Bin marking on this product is a reminder to dispose of the product properly after it has completed its useful life and been removed from service. Metals, plastics, and
other components are recyclable and you can do your part by doing one of the following steps:
• When the equipment is ready to be disposed of, take it to your local or
regional waste collection administration for recycling.
• In some cases, your “end of life” product may be traded in for credit to-
wards the pur-chase of new AMETEK instruments. Contact your dealer to
see if this program is avail-able in your area.
• If you need further assistance in recycling your AMETEK product, contact
our office listed in the front of the instruction manual.
User Manual | 7
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to this product
or any products connected to it.
Use Proper Wiring
To avoid fire hazards, use only the wiring specified in the Installation Chapter of this user’s
manual.
Avoid Electrical Overload
To avoid electrical shock or fire hazard, do not apply a voltage to a terminal that is outside
the range specified for that terminal.
Ground the Product
Follow the grounding instructions provided in the Installation Chapter of this user’s manual.
Before making connections to the input or output terminals of this product, ensure that the
product is properly grounded.
Do Not Operate without Covers
To avoid electric shock or fire hazard, do not operate this product with covers or panels
removed.
Use Proper Fuse
To avoid fire hazard, use only the fuse type and rating specified for this product.
Do Not Operate in Explosive Atmosphere
To avoid injury or fire hazard, do not operate this product in an explosive atmosphere unless
you have purchased options that are specifically designed for these environments.
Product Damage Precautions
Use Proper Power Source
Do not operate this product from a power source that applies more than the voltage specified.
Do Not Operate with Suspected Failures
If you suspect there is damage to this product, have it inspected by qualified service personnel.
Protect From Daylight and Light from Luminaires
Do not install the AMEVision in direct sunlight or light from luminaires as prolonged exposure to these can degrade the sealing properties of the keypad and LCD protective screen.
Integrity of Sealing May Be Impaired by Wear or Deterioration of
Keypad Material
Inspect Keypad Before use.
8 | AMEVision Host Display
WARRANTY AND CLAIMS
We warrant that any equipment of our own manufacture or manufactured for us pursuant to our specifications which shall not be, at the time of shipment thereof by or for us, free from defects in material
or workmanship under normal use and service will be repaired or replaced (at our option) by us free of
charge, provided that written notice of such defect is received by us within twelve (12) months from date
of shipment of portable analyzers or within eighteen (18) months from date of shipment or twelve (12)
months from date of installation of permanent equipment, whichever period is shorter. All equipment requiring repair or replacement under the warranty shall be returned to us at our factory, or at such other location as we may designate, transportation prepaid. We shall examine such returned equipment, and if it is
found to be defective as a result of defective materials or workmanship, it shall be repaired or replaced as
aforesaid. Our obligation does not include the cost of furnishing any labor in connection with the installation of such repaired or replaced equipment or parts thereof, nor does it include the responsibility or cost
of transportation. In addition, instead of repairing or replacing the equipment returned to us as aforesaid,
we may, at our option, take back the defective equipment, and refund in full settlement the purchase price
thereof paid by Buyer.
The warranty shall not apply to any equipment (or part thereof) which has been tampered with or altered
after leaving our control or which has been replaced by anyone except us, or which has been subject to
misuse, neglect, abuse or improper use. Misuse or abuse of the equipment, or any part thereof, shall be
construed to include, but shall not be limited to, damage by negligence, accident, fire or force of the elements. Improper use or misapplications shall be construed to include improper or inadequate protection
against shock, vibration, high or low temperature, overpressure, excess voltage and the like, or operating
the equipment with or in a corrosive, explosive or combustible medium, unless the equipment is specifically designed for such service, or exposure to any other service or environment of greater severity than
that for which the equipment was designed.
The warranty does not apply to used or secondhand equipment nor extend to anyone other than the original purchaser from us.
THIS WARRANTY IS GIVEN AND ACCEPTED IN LIEU OF ALL OTHER WARRANTIES, WHETHER EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION AND WARRANTIES OF FITNESS OR OF MERCHANTABILITY OTHER THAN AS EXPRESSLY SET FORTH HEREIN, AND
OF ALL OTHER OBLIGATIONS OR LIABILITIES ON OUR PART. IN NO EVENT SHALL WE BE
LIABLE UNDER THIS WARRANTY OR ANY OTHER PROVISION OF THIS AGREEMENT FOR
ANY ANTICIPATED OR LOST PROFITS, INCIDENTAL DAMAGES, CONSEQUENTIAL DAMAGES, TIME CHANGES OR ANY OTHER LOSSES INCURRED BY THE ORIGINAL PURCHASER
OR ANY THIRD PARTY IN CONNECTION WITH THE PURCHASE, INSTALLATION, REPAIR
OR OPERATION OF EQUIPMENT, OR ANY PART THEREOF COVERED BY THIS WARRANTY
OR OTHERWISE. WE MAKE NO WARRANTY, EXPRESS OR IMPLIED, INCLUDING WITHOUT
LIMITATION ANY WARRANTIES OF FITNESS OR OF MERCHANTABILITY, AS TO ANY OTHER
MANUFACTURER’S EQUIPMENT, WHETHER SOLD SEPARATELY OR IN CONJUNCTION
WITH EQUIPMENT OF OUR MANUFACTURE. WE DO NOT AUTHORIZE ANY REPRESENTATIVE OR OTHER PERSON TO ASSUME FOR US ANY LIABILITY IN CONNECTION WITH
EQUIPMENT, OR ANY PART THEREOF, COVERED BY THIS WARRANTY.
User Manual | 9
EC Declaration of Conformity
Manufacturer’s Name:
AMETEK Process Instruments
Manufacturer’s Address: 150 Freeport Road
Pittsburgh, PA 15238
Declare under our sole responsibility that the product:
Product Name:
AMEVision Display User Interface
Marking:
Ex nA nC IIC T6 Gc
-20C < Ta < 55C
Conforms to the following standards:
EMC Directive 2004/108/EC:
Immunity:
EN 50082-2: 1995
ELECTROMAGNETIC COMPATIBILITY - GENERIC IMMUNITY STANDARD PART 2. INDUSTRIAL ENVIRONMENT
EN 50082-1: 1997
ELECTROMAGNETIC COMPATIBILITY GENERIC IMMUNITY STANDARD PART 1:
RESIDENTIAL, COMMERCIAL AND LIGHT INDUSTRY
EN 61000-6-2; 1999 ELECTROMAGNETIC COMPATIBILITY (EMC) - PART 6-2: GENERIC STANDARDS
IMMUNITY FOR INDUSTRIAL ENVIRONMENTS
EN 61326; 1997
ELECTRICAL EQUIPMENT FOR MEASUREMENT, CONTROL AND LABORATORY USE
-EMC REQUIREMENTS PART 1: GENERAL REQUIREMENTS
Emmisions:
EN 55011; 1998
INDUSTRIAL, SCIENTIFIC AND MEDICAL EQUIPMENT - RADIO-FREQUENCY DISTURBANCE CHARACTERISTICS - LIMITS AND METHODS OF MEASUREMENT
Low Voltage Directive 73/23/EEC:
EN61010-1; 1993
SAFETY REQUIREMENTS FOR ELECTRICAL EQUIPMENT FOR MEASUREMENT,
CONTROL, AND LABORATORY USE - PART 1: GENERAL REQUIREMENTS
(Amendment A2; 1995)
ATEX Directive 94/9/EC:
EN 60079-0: 2012 EN60079-15: 2010
Certificate Number ITS14ATEX48076X
Notified Body: Intertek
Intertek House, Cleeve Road
Leatherhead, Surrey, KT22 7SB
_____________________________________________________
Robert J. Fiore
Manager of New Product Development
5 May, 2014
10 | AMEVision Host Display
User Manual | 11
INTRODUCTION
Purpose
The AMEVision Display User Interface is a graphical user interface and communications suite for the
WDG-V.
OVERVIEW
Display and Keypad
The AMEVision consists of a 4.2” color, 1/4 VGA display and a 19 key keypad. It
features a web interface, Modbus, trend Data Logging/charting, USB ports for firmware
downloads and parameter/trend data uploads. The host can monitor up to four sensors
simultaneously.
Color 1/4 VGA
Navigation
Buttons
Number Pad
Enter/
Menu Key
Help
Cancel/Escape
AMEVision Display
12 | AMEVision Host Display
Wiring
This wiring section shows you how to make the following connections:
•
AC mains supply wiring to control unit
•
AC mains supply wiring to sensor
•
AMEVision Host Display to sensor
•
AMEVision Host Display to host computer (RS-485 communications)
In addition, this wiring section provides mandatory EMC grounding, shielding, and noise
protection requirements.
General Wiring and Conduit Requirements
This section describes general wiring and conduit requirements.
•
All wiring conductor ratings should be for the minimum temperature required for the
equipment being connected to the analyzer, but not less than 60°C.
•
Use only the applicable ATEX-Certified cable glands to maintain the ATEX rating for
the AMEVision Display User Interface. If not using a conduit entry, leave the factory
ATEX/NEMA-approved plugs intact. Never leave any holes unplugged.
•
Follow all applicable electrical codes for your location.
•
Follow proper grounding, shielding and noise protection practices as described in this
section.
•
Use twisted-pair cable with an overall braided shield, or twisted-pair cable in rigid
metal conduit for all sensor and signal wiring.
•
For AC mains supply wiring, use between 12 and 14 American Wire Gauge (AWG)
or equivalent metric between 3.3 mm 2 and 2.1 mm2. For all signal wiring, use 18 to
22 AWG (0.82 mm2 to 0.33 mm2 ). For sensor connections, Figure 1-02 shows the
recommended sensor cable types for various WDG-style sensors.
•
Use the conduit entry point closest to the connections you are making. Do not add
any additional conduit entry holes!
•
You may use 1” conduit with a 3/4” adapter at the entry to the AMEVision Host Display for signal and sensor wiring.
User Manual | 13
AMEVision Host Display Mains Supply Connections
The AMEVision Host Display can operate using between 85 to 250 volts AC, 47 to 63
Hz. There is no power switch or circuit breaker on the control unit, and it must be protected by installing it on a circuit-protected line, maximum 15 amperes, with a switch or
circuit breaker in close proximity to the control unit and within easy reach of an operator.
Mark the switch or circuit breaker as the control unit disconnecting device.
Mains supply connections to the control unit are as follows:
•
Line - Line connection
•
Neutral - Neutral connection (USA)
•
Chassis Stud - Equipment ground (protective conductor)
Use the 1/2” conduit entry hole for AC mains supply wiring. Use the chassis stud next to
the 1/2” conduit entry hole for equipment ground (protective conductor).
AC (L) and (N) markings are provided by the terminal block for connection of AC power.
These markings are for reference purposes only, such as for use on system wiring diagrams, etc. The system/product has or needs no specific LINE or NEUTRAL connection
for any function, safety or otherwise. The (N) terminal is not internally grounded, nor
needs to be. The system will operate normally regardless of what AC input terminal (L
or N) the AC Line or Neutral is connected to, or, if there is a Neutral used at all (i.e., 208
VAC US power connection).
EMC grounding, shielding, and noise protection
For EMC purposes, under no circumstances should you leave cable shields
disconnected at one end or both ends of the cable (sensor or control unit or
other device).
You must use twisted-pair cable in rigid metal conduit or use twisted pair cable with an
overall braided shield. All cable shields or conduits connecting to the control unit must be
chassis grounded.
14 | AMEVision Host Display
EMC grounding method
Figure 1-3.
•
Shield Ring Method
Connect all shields for that conduit entry (other than power) to a supplied shield terminal ring (see Figure 1-03). This shield ring is a stainless steel ring with a metal tab.
Place the shield ring under the conduit nut. Crimp the shields from all cables for that
conduit entry to a 1/4” female quick disconnect, then push it onto the tab that sticks
out of the conduit shield ring. Keep shields as short as possible.
•
Ground Stud Method
Connect all cable shields for that conduit entry to the grounding stud closest to that
conduit entry hole.
Direct shield grounding methods
User Manual | 15
Transient and RFI interference
This section describes transient and RFI interference precautions:
•
Although there are transient and noise protectors on all control unit I/O connections
(communications, current outputs, sensor, etc.), this protection is intended to act as a
last line of defense against unwanted transient and RFI interference.
•
Proper installation practices to prevent the introduction of transients and noise into
the system must be followed. Inductive loads connected to the control unit must have
transient suppressors installed at the inductive loads. Be sure to place the transient
suppressor as close to the load as possible. Examples of transient suppressors include
MOVs, TRANSORBs, and RC snubbers.
•
AC mains supply wiring should not be run in the same conduit with mains supply
wiring that feeds heavy inductive loads.
•
Avoid running signal wiring in the same cable or conduit with wires that power inductive loads unless all the cables within the conduit are shielded, the inductive loads
are small, and transient suppressors are used at the loads.
•
Do not run signal lines in the same cable or conduit with high voltage lines.
•
For optimum noise protection, control unit mains supply wiring should be connected
to a circuit separate from any circuit that could introduce transients into the system.
As an example, do not run motors, blowers, or air conditioners using the same mains
supply circuit or conduit as the control unit’s mains supply circuit or conduit.
16 | AMEVision Host Display
User Manual | 17
A
B
12.00
304.80
1.44
36.48
CL
1.44
36.48
2
10-15-12
CK
ORIG. ISSUE
OF
10-15-12
CK
ENGR. APPR.
MATERIAL
1
10-15-12
SB
SHEET
10-15-12
SB
DRAWN BY
NTS
2.58
65.53
(3) 3/4" (2.70cm)
NPT CONDUIT
ENTRIES
1
1
ORIGINAL ISSUE
REVISION DESCRIPTION
DATE
10-15-12
SB
DR
SB
APPR
A
SIZE
DWG. NO.
1
203-415
AMEVISION REMOTE INTERFACE
MOUNTING & DIMENSIONS
A
REV
AMETEK
PROCESS & ANALYTICAL INSTRUMENTS DIV.
150 FREEPORT ROAD
PROCESS & ANALYTICAL INSTRUMENTS DIVISION PITTSBURGH, PENNSYLVANIA 15235 USA
(2) 1/2" (2.22cm)
NPT CONDUIT ENTRIES
REPRODUCTION PROHIBITED WITHOUT
PERMISSION BY AMETEK P & AI DIVISION
SCALE
A
REV.
HINGE
SIDE
10.00
254
CONFIDENTIAL
1.44
36.48
2.00
50.80 TYP
DOC. CONT.
TOOL/FIXTURE
IN HOUSE ONLY
REMOVE BURRS AND BREAK SHARP EDGES
1.58
40.13
KEYPAD
9.14
232.19
DISPLAY
1.50
38.10
9.00
228.60
DIMENSIONS ARE IN INCHES
TOLERANCES:
FRACTIONAL 1/32"
ANGULAR: MACH 30'
.010
TWO PLACE DECIMAL
THREE PLACE DECIMAL .005
SURFACE ROUGHNESS 125
THERMOX
PRODUCT LINE
5.97
151.58
2
A
B
AMEVISION DISPLAY USER INTERFACE
Overview
AMEVision Display User Interface provides the capability to easily configure,
calibrate, and monitor up to four (4) WDG-V sensors. A two-wire Modbus RTU
interface connects the AMEVision to each WDG-V sensor. The AMEVision provides the following communication interfaces: Modbus RTU, web interface (TCP/
IP) and USB flash drive capabilities for uploading/downloading data and firmware upgrades.
AMEVision Software Display and Keypad
•
•
•
•
•
4.2” color, 1/4 VGA display
19-key keypad
Web interface (TCP/IP) and MODBUS (RTU)
Trend Data Logging/charting screen
USB port inside for firmware downloads and data uploads
AMEVision Display Screen
The menu display is a two dimensional graphical user interface accessed using the navigation arrows
and ENTER/MENU key. These
keys allow the user to navigate
through the icons and to select the
menu option needed.
AMEVision Keypad
The AMEVision is operated by
pressing the arrows and buttons on
the keypad. ENTER/MENU selects
the option on the screen and CANCEL exits the option. The numeric
keypad is used to enter information
in the fields. Navigation arrows are
used to move throughout the display.
.
18 | AMEVision Host Display
Home Screen
Figure 4-1. Home Screen
The Home screen displays user-defined information and analyzer status. This is
the main source for concentration values, system status and alarms.
1. Gas Concentration
2. Trend Plot (O2 and O2 combustibles only); displays the last 30 minutes.
3. Temperatures (hidden in Normal mode) - use UP arrow to display, DOWN
arrow to hide.
4. Time and Date
5. Sensor Name
6. Last Calibration Date and Time
7. Status Bar - displays analyzer status including Alarms and Warnings
User Manual | 19
When all gases are selected in the Sensor Configuration menu, the Home
screen does not display the Trend plot.
NOTE
When there is an alarm condition, alarm information is displayed on the Status Bar which turns RED for FAIL and YELLOW for WARNINGS.
NOTE
On multi-sensor systems, the left and right arrows navigate through the
connected sensors.
NOTE
NOTE
20 | AMEVision Host Display
The sensor you have selected on the Home screen is the only sensor whose
functions and process will display on the AMEVision menus. To configure or
view additional sensor setups, you must first select that sensor on the Home
screen using the left/right navigation arrows.
Configuration Menu
Figure 4-2. Sensor Configuration menu.
Analog Output
Sensor Configuration
Define parameters that control analog current outputs.
Define the type of sensor you are using and
the gases to be measured.
Alarm Configuration
Define parameters that control the alarm
relays.
Communications
Define AMEVision communication parameters. Select TC/IP or modbus function.
Process Pressure
Select a positive or negative process pressure. Enter process pressure value.
System Time
Set system and display time in 24-hour
format.
Passcode Configuration
Create a system password to limit control
unit access.
Sensor Name Configuration Name sensor and connected sensors in
multiple analyzer configuration.
Selecting Menu Options
Press the ENTER/MENU key on the AMEVision keypad.
Setting Menu Options
Accept the changes you have made by pressing the ENTER/MENU key.
Exiting Menu Options
Press the CANCEL key on the AMEVision keypad.
User Manual | 21
System Passcode
By default, the system passcode is disabled. This menu allows you to create a new
passcode or to disable the passcode requirement.
If you create and enter a passcode, you will be required to enter that passcode
on a lock screen every time you exit the Home screen and before accessing any
of the menu options.
1. From the Main Menu screen, press the ENTER/MENU key to access the Configuration Menu.
2. Use the left and/or right navigation arrows to locate the Passcode Configuration menu and press the ENTER/MENU key.
Figure 4-3. Password Configuration screen.
3. Enter a 5-digit passcode using the numeric keys on the keypad. Press the ENTER/MENU key and enter the passcode again to confirm.
4. Navigate to the ACCEPT button and press the ENTER/MENU key.
NOTE
22 | AMEVision Host Display
To clear an existing passcode or if you do not want to set a passcode, enter
“00000” in the passcode field.
System Name Configuration
NOTE
The default name of the analyzer is WDG-n, where “n” is the address of the
analyzer. If you have multiple sensors, entering a name for each sensor you
want to connect is recommended but not required.
Figure 4-4. System Name Configuration screen.
1. From the Main Menu screen, press the ENTER/MENU key to access the Configuration Menu.
2. Use the left and/or right navigation arrows to locate the System Name Configuration menu and press the ENTER/MENU key.
3. Use the keypad to enter the name(s) of your analyzer(s). Enter the name as
you would on a mobile phone. The keypad displays the characters as you enter them. For example, to select the letter “E”, press the number “3” key twice.
To enter the number “2” press the number “2” key four times “A, B, C, 2.”
Maximum number of characters is 10.
NOTE
4. Navigate to ACCEPT and press the ENTER/MENU key after each name to accept the name of that sensor.
User Manual | 23
System Time
Figure 4-5. System Time screen.
1. From the Main Menu screen, press the ENTER/MENU key to access the Configuration Menu.
2. Use the left and/or right navigation arrows to locate the System Time menu
and press the ENTER/MENU key.
3. Enter the information for your analyzer using a 24-hour time format, i.e.,
noon is 12:00 and 1 PM is 13:00.
4. Navigate to ACCEPT and press the ENTER/MENU key to set the time of that
sensor.
24 | AMEVision Host Display
Communication
1. From the Main Menu screen, press the ENTER/MENU key to access the Configuration Menu.
2. Use the left and/or right navigation arrows to locate the Communicatiuon
menu and press the ENTER/MENU key.
There are two user communication options available: Web Interface (TCP/IP) and
Modbus. Use the left and right arrows to navigate between the two.
TCP/IP (Web Interface) Setup
Figure 4-6. TCP/IP (Web Interface) Configuration screen.
Modbus Communication Setup
Figure 4-7. Modbus Configuration screen.
User Manual | 25
Sensor Configuration
Select Sensor Options
Figure 4-8a. Select Sensor Options screen.
1. From the Main Menu screen, press the ENTER/MENU key to access the Configuration Menu.
2. Use the left and/or right navigation arrows to locate the Sensor Configuration
menu and press the ENTER/MENU key.
3. Options for the various sensor combinations are listed. Specify the options
you are using: Combustibles or Methane or both.
If you have a flow sensor on your analyzer, make sure that you have checked
the “Flow” box on this screen.
NOTE
Figure 4-8b. Select Sensor Options screen. Flow Sensor selection.
NOTE
26 | AMEVision Host Display
Auto Calibration is not a standard feature of the WDG-V Analyzer. Do not
select Auto Calibration unless your analyzer is equipped with a Remote Calibration Unit (RCU).
4. Press the ENTER/MENU key to select each option. To deselect the option,
press ENTER again.
5. Use the arrow keys to navigate between the options. When you have made
your selections, navigate to ACCEPT and press the ENTER/MENU key. Selected options are indicated by an “X” in the check box.
NOTE
The sensor you have selected on the Home display screen is the only sensor
whose functions and process will be displayed on the AMEVision menus. To
view additional sensor setups, you must first select that sensor on the Home
display screen.
Process Pressure
Define your process pressure and whether your process is operating under a
positive or negative pressure. Setting this value corrects oxygen measurements
for the defined pressure.
Figure 4-9. Process Pressure screen.
1. From the Main Menu screen, press the ENTER/MENU key to access the Configuration Menu.
2. Use the left and/or right navigation arrows to locate the Process Pressure
menu and press the ENTER/MENU key.
3. Enter your process pressure.
4. Navigate to the ACCEPT button and press the ENTER/MENU key to accept
your entry.
If you are entering a negative process pressure, use the minus (-) key on the
AMEVision enclosure before entering the number.
NOTE
User Manual | 27
Analog Output
You can direct process readings to any of the three analog output ports.
NOTE
The sensor you have selected on the Home screen is the only sensor whose
functions and process will displayed on this menu. To configure or view any
additional sensor, you must select that sensor on the Home display screen
first.
Figure 4-10. Analog Output configure screen.
Function
Mode
Span
Zero
Track/Hold during Cal**
%O2
Cell mv
Combustibles
Methane
0-20mA
4-20mA
Namur*
Enter reading for 20 mA
(%) for O2 & methane
(ppm) for combustibles
(mV) for cell millivolt
Enter reading for 0/4 mA
(%) for O2 & methane
(ppm) for combustibles
(mV) for cell millivolt
Track during Cal
Hold during Cal
** Define the signals to send to the selected analog output port during a
calibration. You have the option to either hold the last process reading
or track calibration readings..
28 | AMEVision Host Display
* The Namur Standard is implemented as follows:
4-20 ma Signal
Analyzer Condition
0 mA
Analyzer unpowered, or completely failed
3.5 mA
Critical Alarm - analyzer reading unusable
3.8 mA
Reading Under Range (Example - user sets range to 2-10%.
Current reading is 1.9%)
4 to 20 mA
20.5 mA
Normal Operation
Reading Over Range (Example - range is 0-10%. Current
reading is 12%)
1. From the Main Menu screen, press the ENTER/MENU key to access the Configuration Menu.
2. Use the left and/or right navigation arrows to locate the Analog Output menu
and press the ENTER/MENU key.
3. The output that is displayed beneath the sensor name is the active output you
are configuring. Use the navigation arrows to access each of the functions.
Press the ENTER/MENU key to view the drop down menu and use the arrow
keys to make your selection.
4. Enter values using the numeric keypad on the AMEVision enclosure. Use the
period for a decimal point.
5. When you have entered the information for each function, navigate to ACCEPT and press the ENTER/MENU key after each output has been configured
to accept the values. Use the left and right arrows to navigate between Analog
Output 1, 2 and 3.
6. When you have entered the information for each analog output, navigate to
ACCEPT and press the ENTER/MENU key after each output has been configured to accept the values.
7. Press the CANCEL key to return to the Sensor Configuration menu.
User Manual | 29
Alarm Configuration
Figure 4-11. Process Alarm Configuration screen.
Figure 4-12. Function drop down box for alarm configuration.
1. From the Main Menu screen, press the ENTER/MENU key to access the Configuration Menu.
2. Use the left and/or right navigation arrows to locate the Alarm Configuration
menu and press the ENTER/MENU key.
30 | AMEVision Host Display
Alarms: Process, Service, Fault and Data Valid
There are three (3) process relay alarms, (1) fault alarm and (1) data alarm.
Process alarms are activated when the process reading levels, based on what the
user has defined in the setup menus, have been exceeded.
Service alarms activate when there is a hardware malfunction and the analyzer is
not operating correctly.
Data Valid alarms activate when there is a hardware malfunction or when the
output may not be reflective of the process, for example during Warm up, Calibration or Diagnostic modes.
Figure 4-13. Fault alarm configuration screen.
If you set the Alarm Fault by selecting the check box, it will cause all process
alarms (warnings) to trip the service alarm.
User Manual | 31
Calibration
The Calibration Menu allows the user to configure the sensor for calibration, perform the calibration, view the last calibration and calibration history.
Start Calibration
Cal Gas Values
Last Cal Results
Cal History
Auto Cal Config
Flow Calibration
Select gases and cal gas values; start
a calibration
Enter cal gas values
Shows results for the last calibration
Plots the history of gases calibrated
Set Auto Cal timers
Used with flow sensor only
Types of Calibration
AMEVision allows you to perform manual, automatic, or remote calibrations.
Automatic and remote calibrations require a factory-provided remote calibration
unit (RCU) that automatically switches the calibration gases for you; manual calibrations are performed without an RCU, and gases are switched manually.
Manual Calibration
The user defines the calibration gas values and physically switches the gases.
Automatic Calibration
This is an option that allows the system to calibrate itself. The user defines the
calibration gas values, the time each gas will flow through the sensor, and the
frequencies at which automatic calibrations will be performed (for example, every
four hours, every two days, etc.). The system calibrates itself at those predefined
times. Do not select Automatic Calibration unless you have a Remote Calibration
Unit (RCU).
Remote Calibration
The user defines the calibration gas values, but interactively determines how long
each calibration gas should flow from the RCU and into the sensor by pressing
the ENTER/MENU key to accept the values when the gas is stable. The gases,
however, are automatically switched on/off using the RCU. Do not select Remote
Calibration unless you have this option.
NOTE
32 | AMEVision Host Display
Remote and Automatic Calibrations are only available if your analyzer has an
RCU and the ‘AutoCal’ option is enabled in the Sensor Configuration menu.
Cal Gas Values Menu
Use this menu to set and save the calibration gas values for the sensor you
have chosen. These values will be used when calibrating the sensor. This menu
provides a way for the user to select the values before calibrating the analyzer.
However, if the values are not set here, they can still be set during the calibration
process under the Start Calibration menu.
Figure 4-14a.
Cal Gas Values Menu
Figure 4-14b.
Set Cal Gas Values
Figure 4-14c.
Start Calibration Menu
Figure 4-14d.
Calibration Gas Values
1. From the Main Menu screen, press the ENTER/MENU key to access the Calibration Menu.
2. Use the left and/or right navigation arrows to locate the Cal Gas Values menu
and press the ENTER/MENU key.
3. Enter the values for the gases available for the sensor you have designated on
the Home screen.
4. Press the ENTER/MENU key to ACCEPT the values you entered.
User Manual | 33
Start Calibration Menu
Select Gas to be Calibrated
Recovery Duration
Calibration Gas Values
All, Oxygen, Combustibles, Methane
Define a time period for the system to recover from reading calibration gas values
to reading process gases.
Enter the calibration gas values to match the
span and zero calibration gas cylinders you
will use to calibrate or verify your system.
A calibration cannot be performed if there is a Fail alarm active except in the
case of when the alarm is triggered by low flow. Calibration can start or continue during low flow alarm status.
The span gas is the high calibration gas. The zero gas is the low calibration
gas.
NOTE
All gases selected under Sensor Configuration (O2, combustibles, methane)
require calibration but do not have to be calibrated at the same time. However,
calibration will always occur in the same sequence.
1.
2.
3.
4.
O2 Span
Zero
Combustibles Span
Methane Span
For example if the sensor is configured for oxygen, combustibles and methane
and you are skipping the O2 calibration, the order would be as follows:
1. Zero Calibration
2. Combustible Span Calibration
3. Methane Span Calibration
To cancel a calibration at any time, navigate to the ABORT button and press
the ENTER/MENU key.
NOTE
All measurements require a two-point calibration: span and zero, so the zero
will always be performed.
NOTE
NOTE
34 | AMEVision Host Display
Calibrations are only allowed when the sensor is in Normal mode. Check the
Status bar at the bottom of the Home screen if you are unsure of the sensor’s
status.
If you do not select a gas to calibrate, a warning displays indicating that no
gas was selected. Press the CANCEL key to exit the screen.
NOTE
After a calibration is completed and if you have replaced the analyzer cell since
the last calibration, you will see a yellow screen indicating that a cell change
has been detected and asking if you would like to reset the Cell Age. If you
would, navigate to “Yes” and press the ENTER/MENU key. If you would not,
navigate to “No” and press the ENTER/MENU key.
Calibrate the Analyzer
Perform a Manual Calibration
1. From the Main Menu screen, press the ENTER/MENU key to access the Calibration Menu.
2. Use the left and/or right navigation arrows to locate the Start Calibration
menu and press the ENTER/MENU key.
3. You are prompted to select the gases to be calibrated, or choose the “All” option.
4. Enter the Recovery Duration time.
5. Use the right navigation arrow to advance to the next screen.
6. Enter the calibration gas values for the gases you have selected, Use the up
and down navigation arrows to advance to the fields.
NOTE
As a default, any gas values that you entered in the Cal Gas Values menu will
automatically be entered on this screen. You can edit these values and they
will be changed in the Cal Gas Values menu as well.
7. When you have entered all of the values, press the ENTER/MENU key to
START the calibration.
8. Inject the Oxygen Span Gas
-- You are prompted to inject the O2 span gas into the calibration gas inlet port
on the sensor (tubing must be free of oil and dirt). Press the ENTER/MENU
key when you have applied the span gas.
User Manual | 35
9. The screen displays the status of the O2 span gas as “Wait” or “Ready.” When
the status is “Ready,” press the ENTER/MENU key to accept the values.
If you accept the values before the status is “Ready” you will be asked to acknowledge that the reading is not stable before continuing.
10. Inject Zero Gas
--
You are then prompted to apply the O2 zero gas to the calibration inlet port
of the sensor. Press the ENTER/MENU key when you have applied this gas.
11. The system will calibrate the O2 and combustibles zero gas. When the readings have stabilized (Ready), press the ENTER/MENU key to accept the readings.
Figure 4-15. Calibration screen gas status reading.
If the sensor is configured for combustibles:
12. Inject the Combustibles Span Gas
--
You are then prompted to apply the combustibles span gas to the calibration
gas inlet port of the sensor (tubing must be free of oil and dirt). Press the
ENTER/MENU key when you have applied this calibration gas.
13. The system will calibrate the combustibles span gas. Press the ENTER/MENU
key to accept the gas readings.
If the sensor is configured for methane:
14. Inject the Methane Span Gas
--
You are then prompted to apply the methane span gas to the calibration gas
inlet port of the sensor (tubing must be free of oil and dirt). Press the ENTER/
MENU key when you have applied this calibration gas.
15. The system will calibrate the methane span gas. Press the ENTER/MENU key
to accept the gas readings.
36 | AMEVision Host Display
16. When all gases have calibrated, the screen displays the Calibration Results
that include % drift as well as recovery time. If the calibration is successful,
a large green check mark will be displayed. If it is not, a large red X is displayed.
Figure 4-16a. Calibration results
screen. Successful calibration.
Figure 4-16b. Calibration results
screen. Failed calibration.
User Manual | 37
Perform an Automatic Calibration
NOTE
The functions for automatic calibration will only be visible if you have an
RCU and have configured your sensor for Automatic Calibration in the Sensor Configuration menu.
Figure 4-17a. Auto Cal Config Menu
Figure 4-17b. Start Calibration Menu.
There are two ways to perform an automatic calibration:
•
Define the automatic calibration time and date in the Auto-Calibration Configuration (Auto Cal Config) menu. This allows the system to perform calibrations at scheduled times and there is no further user interaction.
•
Use the Start Calibration (or Set Cal Gas) menu to define the gases and values
and then use the Start Calibration menu to perform an immediate calibration.
This requires the user to program the Start Calibration menu and physically
press the ENTER/MENU key to start the calibration. The real-time sequence of
calibration events is displayed on a series of screens which require input from
the user before each step of the calibration continues.
Configuring the Auto Cal using the Auto-Cal Config Menu
1. From the Main Menu screen, press the ENTER/MENU key to access the Calibration Menu.
2. Use the left and/or right navigation arrows to locate the Auto Cal Config
menu and press the ENTER/MENU key.
3. Use the right and left arrows to navigate to the next screen.
38 | AMEVision Host Display
Figure 4-18. Auto-Calibration Configuration screen.
Cal Gas Duration
in Seconds
Define how long you want each calibration gas
to flow through the sensor before switching the
gases. The RCU automatically switches the span
and zero calibration gases based on the calibration gas duration times you specify.
Be sure to allow enough time for the gases to stabilize when you define these times.
Calibration Frequency in Days
Start Time
Hours:Minutes
Start Date
Month:Day:Year
Auto-Calibration
Enabled
Next Calibration
Filled Automatically
You must exit this menu and re-enter to view the next calibration date set by the
system.
You must select the “Enabled” check box.
NOTE
When you have completed this screen, navigate to the ACCEPT button and press
the ENTER/MENU key to continue.
The automatic calibration has now been set and the system will carry out the
calibration per the schedule entered.
Auto Cal using the Calibration Start Menu
1. From the Main Menu screen, press the ENTER/MENU key to access the Calibration Menu.
2. Use the left and/or right navigation arrows to locate the Start Calibration
menu and press the ENTER/MENU key.
3. Navigate to the drop down box and press the ENTER/MENU key. Select “Automatic.”
User Manual | 39
Figure 4-19. Select calibration type.
4. Press the right arrow to continue.
5. Select the gas(es) that you want to calibrate.
6. Enter the time for the Cal Gas Duration in seconds.
7. Enter the time for the Recovery Duration in minutes.
8. Press the right arrow to continue.
NOTE
This is the Calibration Gas Values screen. If you have already entered your
calibration gas values in the Cal Gas Menu, they will be visible in the fields.
Any changes that you make to these values will be reflected in the Cal Gas
Menu as well.
9. When you are satisfied with your values, press the ENTER/MENU key to start
the automatic calibration. The screen displays the Gas Value, Cal Gas and
Time Remaining. To abort the calibration, navigate to the ABORT button and
press the ENTER/MENU key.
NOTE
If O2 is selected for calibration, it will be the first gas calibrated and the zero
gas will be calibrated next, then combustibles and methane. If combustibles
is the first gas selected for calibration, zero gas will be calibrated first, then
combustibles and methane.
10. The screens will advance upon completion of each calibration to the next gas
to be calibrated.
11. When all gases have calibrated, the screen displays the Calibration Results
that include % drift for gases calibrated as well as recovery time. If the calibration is successful, a large green check mark will be on the screen. If it is not, a
large red X is displayed.
40 | AMEVision Host Display
Perform a Remote Calibration
NOTE
The functions for remote calibration will only be visible if you have and RCU
and you have configured your sensor for Automatic Calibration in the Sensor
Configuration menu.
Using the Remote Cal Unit (RCU), the user manually accepts the gas values when
each gas is stable by pressing the ENTER/MENU key, but the RCU switches the
gases on/off automatically.
1. From the Main Menu screen, press the ENTER/MENU key to access the Calibration Menu.
2. Use the left and/or right navigation arrows to locate the Start Calibration
menu and press the ENTER/MENU key.
3. Navigate to the drop down box, press the ENTER/MENU key and choose
“Remote.”
4. Press the right arrow to continue.
5. Select the gas(es) that you want to calibrate.
6. Enter the time for the Recovery Duration in minutes.
7. Press the right arrow to continue.
NOTE
This is the Calibration Gas Values screen. If you have already entered your
calibration gas values in the Cal Gas Menu, they will be visible in the fields.
Any changes that you make to these values will be reflected in the Cal Gas
Menu as well.
8. When you are satisfied with your values, press the ENTER/MENU key to start
the remote calibration. The screen displays the Gas Value, Cal Gas and Status
(Ready/Wait). To abort the calibration, navigate to the ABORT button and
press the ENTER/MENU key.
9. You will be prompted to “Inject the Zero Gas.” Press the ENTER/MENU key
and the RCU will apply the gas.
10. Calibration gas values and status are displayed. Press the ENTER/MENU key
to accept the values.
The status line at the bottom of the screen indicates which gas is being calibrated.
NOTE
User Manual | 41
11. You are then prompted to “Inject the Combustibles Span Gas” Press the ENTER/MENU key to have the RCU inject the gas.
12. Calibration gas values and status are displayed. When Ready, press the ENTER/MENU key to accept the values.
13. When all gases have calibrated, the screen displays the Calibration Results
that include % drift as well as recovery time. If the calibration is successful,
a large green check mark will be displayed. If it is not, a large red X is displayed.
Flow Calibration
You must have a flow sensor on your analyzer to configure flow calibration.
NOTE
Flow sensor calibrations should only be performed on new flow sensors or
when the electronics have been changed. Use the same gas used for O2 span
when calibrating the flow sensor.
NOTE
Make sure that you have checked “Flow” checked on the “Select Sensor Options” submenu under Sensor Configuration.
Figure 4-20a . Flow Calibration should only be done on a new sensor.
1. From the Main Menu screen, press the ENTER/MENU key to access the Calibration Menu.
2. Use the left and/or right navigation arrows to locate the Flow Calibration
menu and press the ENTER/MENU key.
3. Press he ENTER/MENU key to continue with the calibration or press the
CANCEL key to exit the screen.
42 | AMEVision Host Display
Figure 4-20b. Flow Sensor Calibration
Figure 4-20c. Flow Sensor Zero Calibration
4. Turn off the aspirator and calibration gas flow. Press the ENTER/MENU key to
continue or the CANCEL key to abort the operation.
5. The Flow Sensor Zero Calibration screen displays . Press the ENTER/MENU
key to start the zero calibration.
6. The analyzer counts down to zero and when the zero calibration has been
completed, the analyzer will display the Flow Sensor Span Calibration screen.
Figure 4-20d. Flow Sensor Span Calibration Figure 4-20e. Flow Sensor Span Calibration
7. You are prompted to inject the cal gas. Be sure to use aspirator air or zero gas.
Press the ENTER/MENU key to continue and access the Flow Sensor Span
Calibration screen.
8. Press the ENTER/MENU key to start the span calibration. The analyzer counts
down to zero and when the span calibration has been completed, the analyzer will display a results page that will show a RED X if the calibration has
failed or a Green Check if it has passed. Recovery timer counts down to zero.
Press CANCEL to exit.
User Manual | 43
Alarms
Use this menu to view alarms that are currently active and also to view a history
of alarms.
Alarm parameters are set during the configuration process under the Alarm
Configuration menu.
NOTE
1. From the Main Menu screen, press the ENTER/MENU key to access the
Menus.
2. Use the left and/or right navigation arrows to locate the Alarms menu and
press the ENTER/MENU key.
3. Two menu options are visible:
1. Currently Active
2. Alarm History
Green
Red
Yellow
Blue
List of all current alarms
List of status and information changes
Cleared alarms
Fail Status
Warning
Normal
Up to 1000 past events under Alarm History can be viewed with 11 events per page.
NOTE
44 | AMEVision Host Display
For a complete listing of possible alarms, descriptions and possible causes,
please see the Alarm and Status Definitions at the end of this chapter.
Diagnostics
Diagnostics are used to help determine the overall health of the sensor as well as
for basic troubleshooting in the unlikely event that there is a problem with the
analyzer. The display screens give visibility to how the analyzer is running and
where glitches might occur. Diagnostics also provide the ability to test customer
I/O such as analog outputs and relay outputs providing information that is useful
when troubleshooting the interface to the distributed control system (DCS).
1. From the Main Menu screen, press the ENTER/MENU key to access the
Menus.
2. Use the left and/or right navigation arrows to locate the Diagnostics menu
and press the ENTER/MENU key.
There are seven diagnostic menus:
•
Sensor Diagnostics
•
Control Diagnostics
•
Inject Gas
•
Trend Logging
•
Cell Diagnostics
•
Combustible Detector
•
CH4 Detector
User Manual | 45
Sensor Diagnostics
1. Use the left and/or right navigation arrows to locate the Sensor Diagnostics
menu and press the ENTER/MENU key.
Sensor Diagnostics Screen 1
Cell Temperature
Box Temperature
Cell HDC
Box HDC
On Time
Total on Time
degrees Celsius
degrees Celsius
%
%
days, hours
days, hours
2. Use the right navigation arrow on the AMEVision enclosure to move to the
next screen.
Sensor Diagnostics Screen 2
Cell Millivolts
Cell T/C Millivolts
Box RTD Resistance
Electronics Temperature
Analog Out 1 Measured
Flow
Millivolts
Millivolts
Ohms
degrees Celsius
mA
Counts
3. Press the CANCEL key to exit either screen and return to the Diagnostics
menu.
46 | AMEVision Host Display
Control Diagnostics
NOTE
Diagnostics Mode is turned on when changes are made on this menu so that
the system does not accidentally trigger an auto calibration while troubleshooting.
1. Use the left and/or right navigation arrows to locate the Control Diagnostics
menu and press the ENTER/MENU key.
When you press the ENTER/MENU key a message appears on the Status
Bar at the bottom of the screen “Diagnostic Mode: OFF.” This changes to Diagnostic Mode: ON if you make any changes. Press the CANCEL key to exit
and turn off Diagnostics Mode.
2. There are three screens that make up the Control Diagnostics menu: Analog
I/O, Relays, and Close Contact.
Control Diagnostics: Analog I/O Screen
Figure 4-21a. Diagnostic Mode. Set Analog Value screen.
This screen displays the three Outputs available and the Set Values and Actual
Values. You can use the navigation arrows to access the Set Value fields and
change these values.
Making a change to any of these fields will turn on the Diagnostic Mode.
Press the CANCEL key if you want to exit Diagnostics Mode.
3. Use the right navigation arrow to move to the Relays screen.
User Manual | 47
Control Diagnostics: Relays Screen
Figure 4-21b. Diagnostic Mode. Relays screen.
This screen displays the Set Values and Actual Values for the five Outputs and allows the user to Open or Close the output relay.
4. Use the right navigation arrow to move to the Close Contact screen.
Control Diagnostics: Close Contact Screen
Figure 4-21c. Diagnostic Mode. Close Contact screen.
This screen reflects the status of the contact input on the sensor board interface.
The user provides the physical connection to the appropriate terminal block
(Contact In) that allows them to access the switch electronically (open/close) to
trigger a calibration. When Diagnostic Mode is ON, the close contact switch is
open, preventing the analyzer from accidentally triggering an auto calibration.
5. When you press the CANCEL key to exit this menu, a message displays indicating that you are exiting Diagnostics Mode.
48 | AMEVision Host Display
Inject Gas
Use the left and/or right navigation arrows to locate the Inject Gas menu and
press the ENTER/MENU key. Use this menu to test the flow of gas to the analyzer.
1. Use the navigation arrows to select the gas to inject and press the Enter/
MENU key. An “X” appears in the check box of the option selected.
2. Navigate to the START button; press the Enter/MENU key to inject the gas.
Figure 4-22. Diagnostic Mode. Inject Gas screen.
Pressing the ENTER/MENU key to start the test will turn on the Diagnostic
Mode.
3. Press the ENTER/MENU key to end the test and return to the Inject Gas
screen. It will also turn Diagnostic Mode Off.
4. Press the CANCEL key to exit from the menu.
User Manual | 49
Trend Logging
Plotting Data using the Candlestick Format
When plotting the data for trend logging, the AMEVision software uses a “candlestick” method to format the trending. This method allows more data to be
plotted than there is space available on your display giving you more insight into
how your analyzer is operating. It is also a useful tool for the technician who can
see weeks and months of data to determine more easily at what point a problem occurred or to use the data to head off a potential problem. The candlestick
method takes the minimum and maximum of each range of sample data and produces a vertical yellow line - called a candlestick - to represent the sample range’s
minimum and maximum data sets.
Figure 4-23a. Trend data for 25 samples
Figure 4-23b. Trend data compiled with min and max data sets.
Figure 23c. Candlestick data for the same time period.
50 | AMEVision Host Display
The Trend Logging menu allows the user to select a gas and a period of time to
plot and log how the selected gas is trending.
1. Use the left and/or right navigation arrows to locate the Trend Logging menu
and press the ENTER/MENU key.
2. Select the gas to log and press the ENTER/MENU key. An “X” appears in the
check box.
Figure 4-24a. Trend Options Setup menu.
3. Use the right arrow to move to the time period and press the ENTER/MENU
key to select the time period option.
4. Navigate to ACCEPT and press the ENTER/MENU key to start plotting data.
Figure 4-24b. Trend data plot in candlestick format.
5. The data is plotted on the screen.
6. Press the CANCEL key to exit the screen.
User Manual | 51
Cell Diagnostics
This menu displays the information the user needs to determine the health of the
cell and to determine if and when it needs to be replaced.
1. Use the left and/or right navigation arrows to locate the Cell Diagnostics
menu and press the ENTER/MENU key.
2. Cell diagnostics are displayed as in the chart below.
Cell Age
Estimated Cell Life
AT Value
Cell Resistance
Cell mV
Calculated Cell Temp
in days
in months
Ohms
mV
°C
3. Press the CANCEL key to exit this screen.
Combustibles Detector
This menu displays the information the user needs to determine the health of the
combustibles detector and to determine if and when it needs to be replaced.
1. Use the left and/or right navigation arrows to locate the Combustibles Detector menu and press the ENTER/MENU key.
2. Combustibles Detector diagnostics are displayed as in the chart below.
Combustibles Detector Age
Estimated Comb Detector Life
Sensitivity
Comb Active Voltage
Comb Reference Voltage
3. Press the CANCEL key to exit this screen.
52 | AMEVision Host Display
in days
in months
cnts/ppm
volts
volts
CH4 Detector
This menu displays the information the user needs to determine the health of the
methane detector and to determine if and when it needs to be replaced.
1. Use the left and/or right navigation arrows to locate the CH4 Detector menu
and press the ENTER/MENU key.
2. Methane Detector diagnostics are displayed as in the chart below.
CH4 Detector Age
Sensitivity
CH4 Actual Voltage
CH4 Reference Voltage
in days
cnts/%
volts
volts
3. Press the CANCEL key to exit this screen.
User Manual | 53
Utility
Utility Menu
1. From the Main Menu screen, press the ENTER/MENU key to access the
Menus.
2. Use the left and/or right navigation arrows to locate the Utility menu and
press the ENTER/MENU key.
There are five utility menus:
•
Detect Sensor
•
Parameter Management
•
Firmware Upgrade
•
USB Transfer
•
Reset Age
54 | AMEVision Host Display
Detect Sensor
•
Shows currently connected sensors.
•
Provides the ability to scan the network in search of sensors.
1. Use the left and/or right navigation arrows to locate the Detect Sensor menu
and press the ENTER/MENU key.
2. Press the ENTER/MENU key to access the Detect Sensor screen.
Figure 4-25. Detect Sensor screen.
3. The sensors that are connected are automatically displayed on the Sensor
screen. When more than one sensor is connected to the analyzer, a green
indicator designates the enabled sensor and red indicators designate disabled
sensors.
If a sensor is connected but not displayed on the screen, navigate to the
SEARCH button and press the ENTER/MENU key to search for the sensor.
NOTE
4. Use the navigation keys to toggle sensors on/off.
5. Press the CANCEL key to exit the screen.
User Manual | 55
Parameter Management
This menu is used to save or restore sensor user parameters or revert the sensor
parameters to factory default.
1. Use the left and/or right navigation arrows to locate the Parameter Management menu.
2. Press the ENTER/MENU key to access the Sensor Parameters Management
menu. The Sensor Parameter Management screen displays.
3. Use the navigation keys to select the function:
--
SAVE sensor’s current parameters
--
RESTORE previously user-saved sensor parameters or
--
REVERT to the factory default parameters.
4. Press ENTER/MENU to save the user parameters.
5. Press the CANCEL key to exit the screen.
56 | AMEVision Host Display
Firmware Upgrade
This menu displays the currently installed versions of the Sensor and Host firmware. Use this menu to:
•
Verify the installed Sensor Firmware version or download upgraded or new
firmware
•
Verify the installed Host Firmware version or download upgraded or new
firmware
NOTE
Firmware upgrades are carried out using a USB flash drive. It is necessary
to plug the USB into the inside of the AMEVision enclosure to download the
software.
1. Use the left and/or right navigation arrows to locate the Firmware Upgrade
menu and press the ENTER/MENU key.
2. Press the ENTER/MENU key to access the Firmware Upgrade screen.
3. Use the navigation keys to select the option to upgrade and press the ENTER/
MENU key.
4. You will be directed to “Insert the USB drive” containing the option’s firmware. This requires that you open the AMEVision enclosure door and physically connect the USB drive.
5. Using a standard 7/16” wrench loosen and remove the nut of the left side of
the AMEVision enclosure and open the AMEVision enclosure door.
6. Locate the plug-in port for the USB flash drive and plug the drive with the
upgrade software into the port in the AMEVision enclosure. Press the ENTER/
MENU key to continue or the CANCEL key to exit.
Figure 4-26. Inside AMEVision enclosure - USB download port.
User Manual | 57
7. The new updated firmware will download and the display will show “Flashing Sensor (Host) Firmware”
8. The upgrade is successful when the next screen displays a large green check
mark indicating the upgrade was installed.
If the upgrade was not successful, a large red “X” will appear on the display
with the word “Failed!” You will be asked if the USB contains the correct
firmware and if the correct USB was plugged in. Check the USB drive to make
sure it is connected and try again. If you continue to fail to upgrade the firmware, contact Ametek customer service.
NOTE
9. Press the CANCEL key to exit the menu.
USB Transfer
Use this menu to upload data to a USB drive for use on a computer.
Uploads are carried out using a USB flash drive. It is necessary to plug the
USB into the inside of the AMEVision enclosure to upload the data.
NOTE
1. Use the left and/or right navigation arrows to locate the USB Transfer menu
and press the ENTER/MENU key.
2. Press the ENTER/MENU key to access the USB Transfer screen.
3. The user can:
---
Transfer Sensor Parameters
Transfer Trend Data
58 | AMEVision Host Display
Transfer Sensor Parameters
1. Using a standard 7/16” wrench, loosen and remove the nut on the left side of
the AMEVision enclosure to open the AMEVision enclosure door.
2. Locate the plug-in port for the USB flash drive and plug the drive into the
port on the AMEVision enclosure.
3. Navigate to the Transfer Sensor Parameters button and press the ENTER/
MENU key. A screen will display asking if you have plugged in the USB drive.
Navigate to OK and press the ENTER/MENU key.
4. The screen displays: “Transferring Data to USB”
5. The upload is successful when the next screen displays a large green check
mark indicating the Parameter Transfer was successful.
File Name Format for Uploaded Files
Analyzer Name
Month (2)
Day (2)
Year (4)
Hour (24)
Minute (2)
Second (2)
_Interval 1,2,3 (.csv)
WDG1
August
21st
2012
1 PM
27 min
41 sec
Interval 1
WDG1
08
21
2012
13
27
41
_Int1.csv
WDG108212012132741_Int1.csv
NOTE
Interval (1,2, or 3) Three (3) files are stored on the flash drive when you upload files, one for each interval. The filenames end with the specific interval,
i.e., Int1, Int2, Int3).
The extension for uploaded Sensor Parameter files is .csv.
NOTE
If the upload was not successful, a large red “X” will appear on the display
with the word “Failed!” You will be asked if the USB was plugged in. Check
the USB drive to make sure it is connected and try again. If you continue to
fail to upload the sensor parameters, contact Ametek customer service.
6. Press the CANCEL key to exit the menu.
User Manual | 59
Transfer Trend Data
1. Using a standard 7/16” wrench, loosen and remove the nut on the left side of
the AMEVision enclosure to open the AMEVision enclosure door.
2. Locate the plug-in port for the USB flash drive and plug the drive into the
port on the AMEVision enclosure.
3. Navigate to the Transfer Trend Data button and press the ENTER/MENU key.
A screen will display asking if you have plugged in the USB drive. Navigate
to OK and press the ENTER/MENU key.
4. The screen display: “Transferring Data to USB”
5. The upload is successful when the next screen displays a large green check
mark indicating the Trend Transfer was successful.
File Name Format for Uploaded Files
Analyzer Name
Month (2)
Day (2)
Year (4)
Hour (24)
Minute (2)
Second (2)
_Plog 1,2,3 (
WDG1
August
21st
2012
1 PM
27 min
41 sec
plog 1
WDG1
08
21
2012
13
27
41
_plog1.txt
WDG108212012132741_plog1.txt
plog (1,2, or 3) Three (3) files are stored on the flash drive when you upload
trend data files. The filenames end with the specific sequential file, i.e., plog1,
plog2, plog3).
NOTE
The extension for uploaded Trend Data files is .txt.
NOTE
If the upload was not successful, a large red “X” will appear on the display
with the word “Failed!” You will be asked if the USB was plugged in. Check
the USB drive to make sure it is connected and try again. If you continue to
fail to upload the trend data, contact Ametek customer service.
6. Press the CANCEL key to exit the menu.
60 | AMEVision Host Display
Reset Age
This menu allows the user to reset the age of the cell, the combustibles detector
and the CH4 detector.
This should only be done when the device you have selected has changed.
1. Use the left and/or right navigation arrows to locate the Reset Age menu and
press the ENTER/MENU key.
2. Press the ENTER/MENU key to access the Reset Age screen.
3. Use the up and down arrows to navigate to the option whose age you want
to reset and press the ENTER/MENU key.
This will reset the age of your (cell, combustibles detector, CH4 detector). Do
you wish to continue?
4. Navigate to OK and press the ENTER/MENU key to continue, or navigate to
Cancel and press the ENTER/MENU key to cancel.
User Manual | 61
About
This menu option provides the user with information on the Host and Sensors.
1. Use the left and/or right navigation arrows to locate the About menu and
press the ENTER/MENU key.
2. Two menus display:
---
About Host
About Sensor
About Host
1. Navigate to About Host and press the ENTER/MENU key. The information
screen for the Host displays.
---
Host Serial Number
Host Firmware Revision
2. Press the CANCEL key to exit the menu.
About Sensor
1. Navigate to About Sensor and press the ENTER/MENU key. The information
screen for the Sensor displays.
----
Sensor Serial Number
Sensor Hardware Revision
Sensor Firmware Revision
2. Press the CANCEL key to exit the menu.
62 | AMEVision Host Display
ALARM AND STATUS DEFINITIONS
ALARM
DESCRIPTION
TYPE
NOTES
Cell T/C Failure
T/C is not connected or is open
Alarm
Cell Temperature Control
A critical over temperature condition
has occurred
Alarm
Unit must be reset
to clear
Cell Over Temp
Cell temperature is 4 degrees above
the setpoint
Cell Under Temp
Cell temperature is 4 degrees below
the setpoint
Alarm
Only after warm-up
is complete
Cell Temp Rise Failure
Cell heater failed to heat the cell
Alarm
Over Temp Relay Tripped
The cell over-temperature relay
tripped
Alarm
RTD Failure
The Box Temp RTD failed
Alarm
Box Temp High
Box temperature is 4 degrees C
above the setpoint
Alarm
Box Temp Low
Box temperature is 4 degrees C
below the setpoint
Alarm
Box Temp Rise Failure
Box temperature failed to heat
Alarm
Low Sample Flow
Low sample flow detected
Alarm
Cell Failure (open)
Cell is no longer functional
Alarm
Combustible Detector Open
Combustible detector is open or not
connected
Alarm
Only when configured as WDG-VC
Methane Detector Open
CH4 detector is open or not connected
Alarm
Only when configured as WDG-VCM
Cell mV Mismatch
The cell mV measurement does not
match the real time cell measurement
Alarm
Analog Output 1 Error
The analog output measurement
does not match the setting
Alarm
Flow Sensor Failure
The flow sensor failed
Alarm
Cell T/C Measurement Mismatch
The redundant temperature measurement does not match the real
time measurement
Alarm
Unit must be reset
to clear
Only after warm-up
time expires
Only after warm-up
time expires
User Manual | 63
WARNINGS
Cell Life nearing End
The ZrO2 cell is nearing its end of life
Warning
Comb Detector Life nearing
End
The combustible detector sensitivity
is near its end of life
Warning
Methane Detector Life nearing
End
The methane detector sensitivity is
near its end of life
Warning
High Cell mV
The combustible measurement is no
longer valid (set to full scale)
Warning
Analog Output1 Out of Range
The analog output value is saturated
(high or low)
Warning
Analog Ouptut2 Out of Range
The analog output value is saturated
(high or low)
Warning
Analog Output3 Out of Range
The analog output value is saturated
(high or low)
Warning
Last O2 Span Calibration
Failed
The oxygen span calibration failed
Warning
Last O2 Zero Calibration Failed The oxygen zero calibration failed
Warning
Last Comb Calibration Failed
The combustible calibration failed
Warning
Last CH4 Calibration Failed
The methane calibration failed
Warning
Last Flow Calibration Failed
The last flow sensor calibration failed
Warning
Oxygen Calibration Required
O2 Calibration is required
Warning
Combustible Calibration Required
O2 Calibration is required
Warning
Methane Calibration Required
O2 Calibration is required
Warning
64 | AMEVision Host Display
Replacement Parts List
AMEVision HMI Host
Asssy LCD HOST Interface PCB and Guard
7000-330-SE*
Overlay, Keypad, HOST Interface
7000-351-KE
HOST Power Supply, AC/DC, 12V, 40W, Switchmode
1000-616-JE
PCB, Assy, HOST Display, Customer Connection Board
8000-123-SE
* Assembly includes LCD display, display MCU board, battery,
and protective cover
When ordering, provide the serial number of your analyzer to ensure proper parts are
ordered:
AMETEK
Process & Analytical Instruments Division
150 Freeport Road
Pittsburgh, PA, USA 15238
Phone: (412) 828-9040 Fax: (412) 826-0399
User Manual | 65
OPTIONAL BLOW BACK CONFIGURATION
The Blow Back configuration menu can be accessed through the Configuration
menu on the AMEVision controller. For this menu to appear, Blow Back must first
be enabled in the Sensor Configuration menu.
The WDG-V allows the user to choose from 4 blow back trigger types:
1. Manual Only – the user must initiate the blow back manually thru the Utility
Menu. See below.
2. Timed – blow back will occur continuously at a predetermined time interval
(i.e. every 2 hours). It is recommended that the time interval be at least 1 hour
to prevent valve coil overheating.
3. Flow – blow back will automatically be triggered if the flow reading from the
flow sensor drops below a user defined threshold.
4. Timed and Flow – blow back will occur at the predetermined time interval
unless flow drops below the user defined threshold in which case the blow
back will trigger due to low flow.
Low Flow triggered blow back will only occur once per hour.
NOTE
Blowback Interval – Time in hours between automatic timed blow back triggers
Blowback Recovery – Time in seconds when the oxygen and combustible readings are frozen after a blow back has been triggered to allow the system to return
to equilibrium
Low Flow Threshold – Flow reading in % at which a blow back is triggered
66 | AMEVision Host Display
Manual Blow Back Trigger
To manually initiate a blow back,
Utility Menu g Blowback g Trigger A Blowback
Blow Back History
To view the previous 5 blow back events and flow readings, go to Diagnostics
Menu g Blowback History
User Manual | 67
FRENCH WARNINGS /
FRANÇAIS AVERTISSEMENTS
NOTES SUR LA SECURITE
LES AVERTISSEMENTS, LES PRÉCAUTIONS ET LES NOTES CONTENUS DANS
CE MANUEL ATTIRENT L’ATTENTION SUR DES INSTRUCTIONS CRITIQUES
COMME SUIT :
Une procédure de fonctionnement qui n’est pas strictement suivie peut engendrer une blessure de personne ou une contamination de l’environnement.
Une procédure de fonctionnement qui n’est pas strictement suivie peut endommager l’équipement.
Informations importantes à ne pas laisser passer :
NOTE
Sécurité électrique
Une tension jusqu’à 5 kV peut être présente dans le boîtier. Coupez toujours les alimentations
électriques avant d’effectuer une maintenance ou un dépannage. Seuls des techniciens qualifiés électriquement doivent effectuer les connexions électriques et les vérifications de terres.
Une utilisation quelconque de l’équipement de manière non spécifiée par le constructeur
peut compromettre la protection de la sécurité d’origine fournie par l’équipement.
68 | AMEVision Host Display
Mise à la terre
La mise à la terre de l’instrument est obligatoire. Les spécifications de performance et la
protection de sécurité sont invalides si le fonctionnement de l’instrument est effectué depuis
une alimentation électrique à la terre impropre.
Vérifiez la continuité de la terre de tout équipement avant d’alimenter électriquement.
Note importante aux utilisateurs
Il n’y a pas de composant à maintenir à l’intérieur même des composants de l’analyseur
WDG-V. Ne pas retirer le capot d’un analyseur WDG-V. Se référer à du personnel qualifié
pour la maintenance.
Les instructions suivantes s’appliquent à l’unité d’affichage hôte AMEVision :
L’alimentation électrique, les entrées et sorties (E/S) doivent être en accord avec les méthodes
de câblage en zone Class I Division 2 (se référer au code électrique national) et en accord
avec l’autorité locale ayant juridiction.
AVERTISSEMENT – RISQUE D’EXPLOSION – LA SUBSTITUTUON DE
COMPOSANTS PEUT COMPROMETTRE LA CONFORMITE DE ZONE
CLASS I DIVISION 2.
AVERTISSEMENT – RISQUE D’EXPLOSION – EN ZONE EXPLOSIBLE,
COUPEZ L’ALIMENTATION ELECTRIQUE AVANT DE REMPLACER OU
CABLER UN MODULE.
AVERTISSEMENT – RISQUE D’EXPLOSION – NE DECONNECTEZ PAS
L’EQUIPEMENT SAUF SI L’ALIMENTATION ELECTRIQUE A ETE COUPEE OU SI LA ZONE EST SANS RISQUE.
CET EQUIPEMENT EST CONFORME A UNE UTILISATION EN ZONE
CLASS I, DIVISION 2, GROUPES A, B, C ET D OU HORS ZONE EXPLOSIBLE SEULEMENT.
AVERTISSEMENT : L’EXPOSITION A DES PRODUITS CHIMIQUES
PEUT DEGRADER LES PROPRIETES DES JOINTS DES MATERIAUX
UTILISES DU MODULE SUIVANT : MODULE D’ALIMENTATION ELECTRIQUE.
User Manual | 69
AVERTISSEMENT : RISQUE D’EXPLOSION. EN ZONE A RISQUE,
COUPEZ L’ALIMENTATION ELECTRIQUE AVANT DE RETIRER OU DE
REMPLACER UN CABLAGE, UN MODULE, LA BATTERIE OU UN COMPOSANT SAUF S’IL EST CERTAIN QUE LA ZONE NE CONTIENT PAS DE
CONCENTRATIONS DE GAZ OU VAPEUR IMFLAMMABLE.
AVERTISSEMENT : LA SUBSTITUTION DES COMPOSANTS SUIVANTS
PEUT COMPROMETTRE LA CONFORMITE DE ZONE DIVISION 2 DE
L’ALIMENTATION ELECTRIQUE.
AVERTISSEMENT : LA BATTERIE DOIT ETRE REMPLACEE AVEC LA
REFERENCE AMETEK 1000-652-VE.
La température ambiante maximum ce cet analyseur est de 55°C (131°F).
L’afficheur hôte AMEVision est un élément complexe de l’analyseur qui doit être maintenu
par un technicien qualifié avant une expertise en technologie de l’instrument et en systèmes
électriques. AMETEK recommande que tout équipement nécessitant une maintenance soit
renvoyé à l’usine. Vous devriez seulement essayer de réparer ou maintenir cet équipement
après avoir reçu une formation d’un représentant AMETEK / Division P&AI. Si vous décidez
d’effectuer la maintenance de cet analyseur, ayez conscience que des hautes tensions, des
hautes températures et d’autres conditions potentiellement à risque peuvent surgir.
70 | AMEVision Host Display
Sommaire de Sécurité Générale
Lisez attentivement les précautions de sécurité suivantes pour éviter de se blesser et
d’endommager l’équipement ou tout produit connecté à ce dernier.
Utilisez un câblage adéquat
Pour éviter les risques de feu, utilisez seulement le câblage spécifié dans le chapitre Installation de ce manuel de l’utilisateur.
Evitez les surcharges électriques
Pour éviter une électrocution ou un risque de feu, n’alimentez pas électriquement des bornes
qui sont hors de la gamme spécifiée pour ces bornes.
Mettez à la terre l’analyseur
Suivez les instructions de mise à la terre fournies dans le chapitre Installation de ce manuel
de l’utilisateur. Avant de connecter aux bornes d’entrée ou de sortie de l’analyseur, assurezvous que l’analyseur est correctement mis à la terre.
Ne pas faire fonctionner sans capot
Pour éviter une électrocution ou un risque de feu, ne faites pas fonctionner l’analyseur sans
capot ou avec un panneau retiré.
Utilisez des fusibles adéquats
Pour éviter tout risque de feu, utilisez des fusibles de types et classes spécifiés pour cet
analyseur.
Ne faites pas fonctionner en atmosphère explosible
Pour éviter toute blessure ou risque de feu, ne faites pas fonctionner l’analyseur dans une
atmosphère explosible, sauf si vous avez acheté les options spécifiquement conçues pour
ces environnements.
PRECAUTIONS AUX DOMMAGES DE L’ANALYSEUR
Utilisez une alimentation électrique adéquate
Ne faites pas fonctionner l’analyseur avec une alimentation électrique qui
fournit plus que la tension électrique spécifiée.
Ne faites pas fonctionner avec les défaillances suspectes
Si vous suspectez que l’analyseur est endommagé, demandez à un technicien
qualifié de l’inspecter.
User Manual | 71
Mise à la terre CEM (EMC), blindage et protection contre le bruit
Dans un but de compatibilité électromagnétique, en aucune circonstance
vous ne devez laisser l’écran d’un câble déconnecté en une ou deux extrémités (unité de contrôle ou de terrain ou autre appareil).
72 | AMEVision Host Display