Teledyne Analytical Instruments OPERATING INSTRUCTIONS FOR Model 2020 Thermal Conductivity Analyzer
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Thermal Conductivity Analyzer
OPERATING INSTRUCTIONS FOR
Model 2020
Thermal Conductivity Analyzer
DANGER
HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING
SYSTEM.
PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.
HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST
FOR A TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.
ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING. BEFORE
CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/
MANAGER.
Teledyne Analytical Instruments
P/N M67677
08/06/1999
ECO # 99-0323
i
Model 2020
Copyright © 1999 Teledyne Analytical Instruments
All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer
language in whole or in part, in any form or by any means, whether it be electronic,
mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of
Teledyne Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91749-1580.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us free
from defects of material and of construction, and that our liability shall be limited to
replacing or repairing at our factory (without charge, except for transportation), or at
customer plant at our option, any material or construction in which defects become
apparent within one year from the date of shipment, except in cases where quotations or
acknowledgments provide for a shorter period. Components manufactured by others bear
the warranty of their manufacturer. This warranty does not cover defects caused by wear,
accident, misuse, neglect or repairs other than those performed by Teledyne or an authorized service center. We assume no liability for direct or indirect damages of any kind and
the purchaser by the acceptance of the equipment will assume all liability for any damage
which may result from its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our
apparatus, and to make any alterations in the dimensions, shape or weight of any parts, in
so far as such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its user, and serves as a tool by
which valuable data can be gathered. The information provided by the instrument may
assist the user in eliminating potential hazards caused by his process; however, it is
essential that all personnel involved in the use of the instrument or its interface, with the
process being measured, be properly trained in the process itself, as well as all instrumentation related to it.
The safety of personnel is ultimately the responsibility of those who control process
conditions. While this instrument may be able to provide early warning of imminent danger,
it has no control over process conditions, and it can be misused. In particular, any alarm or
control systems installed must be tested and understood, both as to how they operate and
as to how they can be defeated. Any safeguards required such as locks, labels, or redundancy, must be provided by the user or specifically requested of Teledyne at the time the
order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions. The
purchaser is responsible for the training of personnel, for providing hazard warning
methods and instrumentation per the appropriate standards, and for ensuring that hazard
warning devices and instrumentation are maintained and operated properly.
Teledyne Analytical Instruments, the manufacturer of this instrument, cannot
accept responsibility for conditions beyond its knowledge and control. No statement
expressed or implied by this document or any information disseminated by the manufacturer or its agents, is to be construed as a warranty of adequate safety control under the
user’s process conditions.
ii
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Specific Model Information
The instrument for which this manual was supplied may incorporate one or more
options not supplied in the standard instrument. Commonly available options are listed
below, with check boxes. Any that are incorporated in the instrument for which this
manual is supplied are indicated by a check mark in the box.
Instrument Serial Number: _______________________
Standard Options Included in the Instrument with the Above Serial Number:
q 2020L:
Gas selector panel consisting of sample/ref flow meters with
stainless steel control valves, tubing and fittings.
q 2020C:
Auto Calibration valves (zero/span) built-in gas selector panel
and control valves are electronically controlled to provide
synchronization with the analyzer’s operations.
q 2020R:
Sealed reference TC cell (application dependent, contact
factory).
Special Options:
q 2020F:
Groups C & D Flame Arrestors with Flow Control Gas Panel.
q 2020H:
Stainless Cell Block with Gold Filaments.
q 2020O:
Groups B Flame Arrestors with Flow Control Gas Panel.
q 2020P:
Groups C & D Flame Arrestors with Cal Valves and Flow
Control Gas Panel.
q 2020Q:
Groups B Flame Arrestors with Cal Valves and Flow Control
Gas Panel.
Teledyne Analytical Instruments
iii
Model 2020
Table of Contents
1 Introduction
1.1
1.2
1.3
1.4
1.5
Overview ........................................................................ 1-1
Typical Applications ....................................................... 1-1
Main Features of the Analyzer ....................................... 1-2
Model Designations ....................................................... 1-3
Operator Interface (Front Panel) .................................... 1-3
1.5.1 UP/DOWN Switch ................................................ 1-4
1.5.2 ESCAPE/ENTER Switch ..................................... 1-5
1.6 Recognizing Difference Between LCD & VFD ............... 1-5
1.7 Equipment Interface (Rear Panel).................................. 1-5
1.8 Gas Connections ........................................................... 1-6
2 Operational Theory
2.1 Introduction .................................................................... 2-1
2.2 Sensor Theory ............................................................... 2-1
2.2.1 Sensor Configuration .............................................. 2-1
2.2.2 Calibration ............................................................... 2-2
2.2.3 Effects of Flowrate and Gas Density ....................... 2-3
2.2.4 Measurement Results ............................................. 2-3
2.3 Electronics and Signal Processing ................................ 2-3
2.4 Temperature Control ...................................................... 2-5
3 Installation
3.1 Unpacking the Analyzer ................................................. 3-1
3.2 Mounting the Analyzer ................................................... 3-1
3.3 Electrical Connections (Rear Panel) .............................. 3-3
3.3.1 Primary Input Power .............................................. 3-3
3.3.2 Fuse Installation..................................................... 3-4
3.3.3 Voltage Selections ................................................. 3-4
3.3.4 Analog Outputs ...................................................... 3-4
3.3.5 Alarm Relays ......................................................... 3-6
3.3.6 Digital Remote Cal Inputs ...................................... 3-7
3.3.7 Range ID Relays .................................................... 3-8
3.3.8 Network I/O ............................................................ 3-9
3.3.9 RS-232 Port ........................................................... 3-9
3.3.10 Remote Probe Connector ...................................... 3-10
3.4 Gas Connections ........................................................... 3-11
3.4.1 Sample System Design ......................................... 3-13
3.4.2 Pressure and Flow Rate Regulation ...................... 3-14
iv
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
3.4.3 VENT Exhaust ....................................................... 3-14
3.4.4 SAMPLE Gas......................................................... 3-15
3.4.5 REFERENCE Gas ................................................. 3-15
3.4.6 ZERO Gas ............................................................. 3-16
3.4.7 SPAN Gas .............................................................. 3-16
3.5 Testing the System ........................................................ 3-16
4 Operation
4.1 Introduction .................................................................... 4-1
4.2 Using the Data Entry and Function Buttons ................... 4-2
4.2.1 Mode/Function Selection ....................................... 4-2
4.2.1.1 Analysis Mode ............................................... 4-2
4.2.1.2 Setup Mode ................................................... 4-4
4.2.2 Data Entry .............................................................. 4-5
4.2.2.1 ENTER .......................................................... 4-5
4.2.2.2 Escape ........................................................... 4-5
4.3 The System Function ..................................................... 4-6
4.3.1 Setting the Display ................................................. 4-6
4.3.2 Setting up an Auto-Cal ........................................... 4-6
4.3.3 Password Protection .............................................. 4-7
4.3.3.1 Entering the Password ................................... 4-7
4.3.3.2 Installing or Changing the Password ............. 4-8
4.3.4 Logging Out ........................................................... 4-10
4.3.5 System Self-Diagnostic Test .................................. 4-10
4.3.6 The Model Screen ................................................. 4-11
4.3.7 Checking Linearity with Algorithm .......................... 4-11
4.4 The Zero and Span Functions ....................................... 4-12
4.4.1 Zero Cal ................................................................. 4-13
4.4.1.1 Auto Mode Zeroing ........................................ 4-13
4.4.1.2 Manual Mode Zeroing .................................... 4-14
4.4.1.3 Cell Failure ..................................................... 4-15
4.4.2 Span Cal ................................................................ 4-16
4.4.2.1 Auto Mode Spanning ..................................... 4-16
4.4.2.2 Manual Mode Spanning ................................. 4-17
4.5 The Alarms Function ...................................................... 4-17
4.6 The Range Select Function ........................................... 4-19
4.6.1 Manual (Select/Define Range) Screen .................. 4-20
4.6.2 Auto (Single Application) Screen ........................... 4-20
4.6.3 Precautions ............................................................ 4-22
4.7 The Analyze Function .................................................... 4-22
4.8 Programming ................................................................. 4-23
Teledyne Analytical Instruments
v
Model 2020
4.8.1 The Set Application Screen ................................... 4.24
4.8.2 The Curve Algorithm Screen ................................. 4-26
4.8.2.1 Checking the Linearization ............................ 4-26
4.8.2.2 Manual Mode Linearization............................ 4-27
4.8.2.3 Auto Mode Linearization ................................ 4-28
4.9 Special Function Setup .................................................. 4-29
4.9.1 Output Signal Reversal .......................................... 4.29
4.9.1.1 Output Signal Reversal .................................. 4-29
4.9.1.2 Output Signal Offset ...................................... 4-30
4.9.2 Polarity Reversal .................................................... 4-30
4.9.3 Gain Preset ............................................................ 4.31
Maintenance
5.1 Routine Maintenance ..................................................... 5-1
5.2 System Self Diagnostic Test........................................... 5-1
5.3 Fuse Replacement ......................................................... 5-2
5.4 Major Internal Components ........................................... 5-3
5.5 Voltage Selections ......................................................... 5-3
5.6 Cell, Heater, or Thermistor Replacement ....................... 5-5
5.6.1 Removing the Cell Compartment ........................... 5-5
5.6.2 Removing and Replacing the Cell Block ................ 5-6
5.6.3 Removing the Heater and/or Thermocouple .......... 5-7
5.6.4 Replacing the Heater and/or Thermocouple .......... 5-8
5.7 Cleaning ......................................................................... 5-9
5.8 Phone Numbers ............................................................. 5-9
Appendix
A-1 Specifications ................................................................. A-1
A-2 Recommended 2-Year Spare Parts List ......................... A-3
A-3 Drawing List ................................................................... A-4
vi
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Introduction 1
Introduction
1.1
Overview
The Analytical Instruments Model 2020 Thermal Conductivity Analyzer, explosion proof, UL and CSA listed for class 1, DIV 1, Groups B, C,
and D service, is a versatile microprocessor-based instrument for measuring a component gas in a background gas, or in a specific mixture of
background gases. It compares the thermal conductivity of a sample stream
with that of a reference gas of known composition. The 2020 can—
• measure the concentration of one gas in a mixture of two gases.
• measure the concentration of a gas in a specific mixture of
background gases.
• measure the purity of a sample stream containing a single
impurity or a mixture of impurities.
The standard 2020 is preprogrammed with automatic linearization
algorithms for a large number of gases and gas mixtures. The factory can
add to this data base for custom applications, and the sophisticated user can
add his own unique applications.
Many of the Model 2020 features covered in this manual are optional,
selected according to the customers specific application. Therefore, the user
may find much here that does not apply to his instrument. This is unavoidable due to the number of possible combinations of features available. We
have endeavored to make the manual as usable and convenient as possible,
in light of this flexibility.
Teledyne Analytical Instruments
1-1
Model 2020
1 Introduction
1.2
Typical Applications
A few typical applications of the Model 2020 are:
• Power Generation
• Air liquefaction
• Chemical reaction monitoring
• Steel manufacturing and heat treating
• Petrochemical process control
• Quality assurance
• Refrigeration and storage
• Gas proportioning control.
1.3
Main Features of the Analyzer
The main features of the Model 2020 Thermal Conductivity Analyzer
include:
1-2
•
Three independent, user definable, analysis ranges allow up to
three different gas applications with one concentration range
each, or up to three concentration ranges for a single gas application, or any combination.
•
Special recalibration range for multiple applications. Recalibrating one, recalibrates all.
•
Automatic, independent linearization for each range.
•
Auto Ranging allows analyzer to automatically select the proper
preset range for a given single application. Manual override
allows the user to lock onto a specific range of interest.
•
RS-232 serial digital port for use with a computer or other
digital communications device.
•
Six adjustable set points concentration with two alarms and a
system failure alarm relays.
•
Extensive self-diagnostic testing, at startup and on demand.
•
Sample and Hold for holding analyzer’s output during Auto
calibration mode.
•
A 2-line alphanumeric display screen, driven by microprocessor
electronics, that continuously prompts and informs the operator.
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Introduction 1
•
High resolution, accurate indication of target or impurity gas
concentration from large, bright, meter readout. (0-9999 ppm
through 0-100 % depending on types of gas involved.)
•
Standard, proven sensor cell design.
•
Wide range of custom applications, ranges, and linearization.
•
Microprocessor based electronics: 8-bit CMOS microprocessor
with 32 kB RAM and 128 kB ROM.
•
Auto and remote calibration capabilities.
•
Four analog outputs: two for measurement (0–1 V dc and
Isolated 4–20 mA dc) and two for range identification.
•
Compact and versatile design: Small footprint, yet internal
components are accessible.
1.4
Model Designations
The Model 2020 is ordinarily custom programmed at the factory to fit
the customer’s application. Many parameters, including the number of
channels, the gas application, the materials specification of the sampling
system, and others, are options. The most common options, are covered in
this manual. See the Specific Model Information checklist in the front
matter of this manual for those that apply to your Model 2020 analyzer.
Some standard models that are not covered in this manual are listed here.
Models 2000B:
NEMA-4, bulkhead mounted enclosure for general
purpose, nonhazardous environments.
Models 2010:
Split architecture models using a sealed explosion-proof
enclosure for the Analysis Unit and a general purpose
remote Control Unit for installation in a safe area.
Models 2020:
Both the analysis section and control unit are in a single
explosion proof enclosure.
1.5
Operator Interface (Front Panel)
The Model 2020 is housed in a explosion proof housing. See Figure 11. The front panel has two single operator controls, a digital meter, and an
alphanumeric display. They are described briefly here and in detail in the
Operations chapter of this manual.
Teledyne Analytical Instruments
1-3
Model 2020
1 Introduction
Figure 1-1: Model 2020 Front Panel
1.5.1
UP/DOWN Switch
Functions: The UP/DOWN switch is used to select the function to be
performed. Choose UP or DOWN to scroll through the following list of
fourteen functions:
• AUTO-CAL Set up an automatic calibration sequence.
• PSWD
Install a password to protect your analyzer setup.
• LOGOUT
Locks Setup Mode.
• MODEL
Displays model and version of analyzer.
• SELF-TEST Runs internal diagnostic program, displays results.
1-4
• SPAN
Span calibrate the analyzer.
• ZERO
Zero calibrate the analyzer.
• ALARMS
Set the alarm setpoints and attributes.
• RANGE
Set up the 3 user definable ranges for the instrument.
• APPLICATION
Set up the 3 definable application ranges
• ALOGORITHM
Set up the linearization
• CAL-INDEPD
Calibration range independently
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Introduction 1
• CONTRAST Adjust LCD contrast. Contrast Function is DISABLED
(Refer to Section 1.6)
• STANDBY
Leave analyzer powered, but no outputs or displays.
WARNING: THE POWER CABLE MUST BE DISCONNECTED TO
FULLY REMOVE POWER FROM THE INSTRUMENT.
Subfunctions: Once a Function is entered, the UP/DOWN switch is
used to select between any subfunctions displayed on the VFD screen.
Parameter values: When modifiable values are displayed on the
VFD, the UP/DOWN switch can be used to increment or decrement the
values.
1.5.2
ESCAPE/ENTER Switch
Data Entry: The ESCAPE/ENTER switch is used to input data, from
the alphanumeric VFD screen into the instrument:
•
Escape Moves VFD display back to the previous screen in a
series. If none remains, returns to the Analyze screen.
With subfunction selected, moves VFD back through
items on screen, to first item, then moves VFD to
previous display.
•
Enter
With a Subfunction Selected: Moves VFD on to the
next screen in a series. If none remains, returns to the
Analyze screen.
With a Value Selected: Enters the value into the
analyzer as data. Advances VFD to next operation.
(See Chapter 4 for details.)
1.6 Recognizing Difference Between LCD & VFD
LCD has GREEN background with BLACK characters. VFD has
DARK background with GREEN characters. In the case of VFD - NO
CONTRAST ADJUSTMENT IS NEEDED.
1.7
Equipment Interface
The electrical connection are described briefly here and in detail in
chapter 3, Installation.
Electrical Connections: The electrical connections on the electrical
connector panel are described briefly here, and in more detail in chapter 3
Installation.
Teledyne Analytical Instruments
1-5
Model 2020
1 Introduction
•
Power Connection
115 or 230 V dc, 50 or 60 Hz.
•
Analog Outputs
0-1 V dc concentration plus 0-1 V dc
range ID. Additional, isolated 4-20 mA
dc plus 4-20 mA dc range ID available.
•
Alarm Connections
2 concentration alarms and 1 system
alarm.
•
RS-232 Port
Serial digital concentration signal
output and control input.
•
Remote Valves
Used for controlling external solenoid
valves, if desired.
•
Remote Sensor
Used for external sensor and
thermocouple, if desired.
•
Remote Span/Zero
Digital inputs allow external control of
analyzer calibration.
•
Calibration Contact
To notify external equipment that
instrument is being calibrated and
readings are not monitoring sample.
•
Range ID Contacts
Four separate, dedicated, range relay
contacts. Low, Medium, High, Cal.
•
Network I/O
Serial digital communications for local
network access. For future expansion.
Not implemented at this printing.
1.8
Gas Connections
The gas connectors are on the bottom of the Model 2020 chassis near
the front doorl.
A sample system must be provided for introduction of zero and span
gas, as well as sample gas, into the sample path, and for controlling the
flowrates through the sample and reference paths of the analyzer. Appropriate pressure reducing regulators must be installed at all gas supply sources.
Gas Connector-and-Selector Panels for specific applications are
available at additional cost. These panels are optional designed to substitute
a standard front panel.
For those customers wishing to incorporate their own sample controls,
the recommended system piping schematic is included among the drawings
at the rear of the manual.
1-6
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Operational Theory 2
Operational Theory
2.1
Introduction
The analyzer is composed of two subsystems:
1. Thermal Conductivity Sensor
2. Electronic Signal Processing, Display and Control.
The sensor is a thermal conductivity comparator that continuously
compares the thermal conductivity of the sample gas with that of a reference gas having a known conductivity.
The electronic signal processing, display and control subsystem
simplifies operation of the analyzer and accurately processes the sampled
data. A microprocessor controls all signal processing, input/output, and
display functions for the analyzer.
2.2
Sensor Theory
For greater clarity, Figure 2-1 presents two different illustrations, (a)
and (b), of the operating principle of the thermal conductivity cell.
2.2.1 Sensor Configuration
The thermal conductivity sensor contains two chambers, one for the
reference gas of known conductivity and one for the sample gas. Each
chamber contains a pair of heated filaments. Depending on its thermal
conductivity, each of the gases conducts a quantity of heat away from the
filaments in its chamber. See Figure 2-1(a).
The resistance of the filaments depends on their temperature. These
filaments are parts of the two legs of a Wheatstone bridge circuit that
unbalances if the resistances of its two legs do not match. See Figure
2-1(b).
Teledyne Analytical Instruments
2-1
2 Operational Theory
Model 2020
Figure 2-1: Thermal Conductivity Cell Operating Principle
If the thermal conductivities of the gases in the two chambers are
different, the Wheatstone bridge circuit unbalances, causing a current to
flow in its detector circuit. The amount of this current can be an indication
of the amount of impurity in the sample gas, or even an indication of the
type of gas, depending on the known properties of the reference and
sample gases.
The temperature of the measuring cell is regulated to within 0.1 °C by
a sophisticated control circuit. Temperature control is precise enough to
compensate for diurnal effects in the output over the operating ranges of
the analyzer. (See Specifications in the Appendix for details.)
2.2.2 Calibration
Because analysis by thermal conductivity is not an absolute measurement, calibration gases of known composition are required to fix the
upper and lower parameters (“zero” and “span”) of the range, or ranges, of
analysis. These gases must be used periodically, to check the accuracy of
the analyzer.
During calibration, the bridge circuit is balanced, with zero gas
against the reference gas, at one end of the measurement range; and it is
sensitized with span gas against the reference gas at the other end of the
measurement range. The resulting electrical signals are processed by the
analyzer electronics to produce a standard 0-1V, or an isolated 4–20 mA
dc, output signal, as described in the next section.
2-2
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Operational Theory 2
2.2.3 Effects of Flowrate and Gas Density
Because the flowrate of the gases in the chambers affects their cooling
of the heated filaments, the flowrate in the chambers must be kept as equal,
constant, and low as possible.
When setting the sample and reference flowrate, note that gases
lighter than air will have an actual flowrate higher than indicated on the
flowmeter, while gases heavier than air will have an actual flowrate lower
than indicated. Due to the wide range of gases that are measured with the
Thermal Conductivity Analyzer, the densities of the gases being handled
may vary considerably.
Then, there are limited applications where the reference gas is in a
sealed chamber and does not flow at all. These effects must be taken in
consideration by the user when setting up an analysis.
2.2.4 Measurement Results
Thermal conductivity measurements are nonspecific by nature. This
fact imposes certain limitations and requirements. If the user intends to
employ the analyzer to detect a specific component in a sample stream, the
sample must be composed of the component of interest and one other gas
(or specific, and constant, mixture of gases) in order for the measured
heat-transfer differences to be nonambiguous.
If, on the other hand, the user is primarily interested in the purity of a
process stream, and does not require specific identification of the impurity,
the analyzer can be used on more complex mixtures.
2.3
Electronics and Signal Processing
The Model 2020 Thermal Conductivity Analyzer uses an 8031 microcontroller, Central Processing Unit—(CPU) with 32 kB of RAM and 128
kB of ROM to control all signal processing, input/output, and display
functions for the analyzer. System power is supplied from a universal
power supply module designed to be compatible with any international
power source. (See Major Internal Components in chapter 5 Maintenance
for the location of the power supply and the main electronic PC boards.)
The Temperature Control board is mounted under the electrical
connection board.. The signal processing electronics including the microprocessor, analog to digital, and digital to analog converters are located on
the Motherboard at the front door of the unit. The Preamplifier board is
Teledyne Analytical Instruments
2-3
2 Operational Theory
mounted on top of the Motherboard as shown in the figure 5.4. These
boards are accessible after removing the back panel. Figure 2-2 is a block
diagram of the Analyzer electronics.
Figure 2-2: Block Diagram of the Model 2020 Electronics
2-4
Teledyne Analytical Instruments
Model 2020
Thermal Conductivity Analyzer
Operational Theory 2
The Temperature Control keeps the temperature of the measuring cell
regulated to within 0.1 degree C. A thermistor is used to measure the
temperature, and a zero-crossing switch regulates the power in a cartridgetype heater. The result is a sensor output signal that is temperature independent.
In the presence of dissimilar gases the sensor generates a differential
voltage across its output terminals. A differential amplifier converts this
signal to a unipolar signal, which is amplified in the second stage, variable
gain amplifier, which provides automatic range switching under control of
the CPU. The output from the variable gain amplifier is sent to an 18 bit
analog to digital converter.
The digital concentration signal along with input from the Gas Selector Panel is processed by the CPU and passed on to the 12-bit DAC, which
outputs 0-1 V dc Concentration and Range ID signals. An voltage-tocurrent converter provides 4-20 mA dc concentration signal and range ID
outputs.
The CPU also provides appropriate control signals to the Displays,
Alarms, and External Valve Controls, and accepts digital inputs for external Remote Zero and Remote Span commands. It monitors the power
supply through an analog to digital converter as part of the data for the
system failure alarm.
The RS-232 port provides two-way serial digital communications to
and from the CPU. These, and all of the above electrical interface signals
are described in detail in chapter 3 Installation.
2.4. Temperature Control
For accurate analysis the sensor of this instrument is temperature
controlled to 60oC.
Teledyne Analytical Instruments
2-5
2 Operational Theory
2-6
Teledyne Analytical Instruments
Model 2020
Thermal Conductivity Analyzer
Installation 3
Installation
Installation of the Model 2020 Analyzer includes:
1. Unpacking
2. Mounting
3. Gas connections
4. Electrical connections
5. Testing the system.
3.1
Unpacking the Analyzer
The analyzer is shipped ready to install and prepare for operation.
Carefully unpack the analyzer and inspect it for damage. Immediately
report any damage to the shipping agent.
The four gas fittings that mate with the 1/4 NPT gas ports on the
Model 2020, are not included. They must be supplied by the customer.
3.2
Mounting the Analyzer
The Model 2020 is designed for bulkhead mounting in hazardous
environments. There are four mounting lugs—one in each corner of the
enclosure, as shown in Figure 3-1. The outline drawing, at the back of this
manual, gives the mounting hole size and spacing. The drawing also contains the overall dimensions. Do not forget to allow an extra 13/8" for the
hinges.
Be sure to allow enough space in front of the enclosure to swing the
door open—a 16 1/4" radius, as shown in Figure 3-2.
All electrical connections are made via cables which enter the explosion-proof housing through ports in its side. No conduit fittings are supplied. The installer must provide two 3/4" NPT and two 1" NPT adapters
and the appropriate sealing conduit.
Teledyne Analytical Instruments
3-1
3 Installation
Model 2020
Hinge
Figure 3-1a: Internal Views of the Model 2020
H
3-2
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Installation 3
Figure 3-2: Required Front Door Clearance
3.3
Electrical Connections
Figure 3-3 shows the Model 2020 Electrical Connector Panel. There
are terminal blocks for connecting power, communications, and both
digital and analog concentration outputs.
For safe connections, ensure that no uninsulated wire extends outside
of the connectors they are attached to. Stripped wire ends must insert
completely into terminal blocks. No uninsulated wiring should be able to
come in contact with fingers, tools or clothing during normal operation.
3.3.1 Primary Input Power
The power cord receptacle and fuse block are located in the same
assembly. Insert the female plug end of the power cord into the power cord
receptacle.
DANGER: POWER IS APPLIED TO THE INSTRUMENT'S CIRCUITRY AS LONG AS THE INSTRUMENT IS CONNECTED TO THE POWER SOURCE. THE STANDBY
FUNCTION IS FOR SWITCHING POWER ON / OFF
TO THE DISPLAY AND OUTPUTS ONLY.
The standard power supply requires a 115 V ac, 50-60 Hz power
source. If you have the -N option, you will require 220 V ac, 50-60 Hz
power.
Teledyne Analytical Instruments
3-3
3 Installation
Model 2020
3.3.2 Fuse Installation
The fuse block, at the right of the power cord receptacle, accepts US
or European size fuses. A jumper replaces the fuse in whichever fuse
receptacle is not used. Be sure to install the proper fuse as part of installation. (See Fuse Replacement in chapter 5, maintenance.)
3.3.3 Voltage Selections
There is a switch on the interface board, inside the instrument, that
selects the working voltage between 230/115 VAC.
230V
115V
Voltage Selector Switch
Make sure the switch is in the proper position before
powering the instrument.
3.3.4 Analog Outputs
There are four DC output signal connectors on the panel. There are
two wires per output with the polarity noted. See Figure 3-4. The outputs
are:
0–1 V dc % of Range: Voltage rises linearly with increasing concentration, from 0 V at 0 concentration to 1 V at full
scale. (Full scale = 100% of programmable range.)
0–1 V dc Range ID:
0.25 V = Range 1, 0.5 V = Range 2, 0.75 V =
Range 3, 1 V = Cal Range.
4–20 mA dc % Range: Current rises linearly with concentration, from 4
mA at 0 concentration to 20 mA at full scale. (Full
scale = 100% of programmable range.)
4–20 mA dc Range ID: 8 mA = Range 1, 12 mA = Range 2, 16 mA =
Range 3, 20 mA = Range 4.
3-4
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Installation 3
Figure 3-4: Analog Output Connections
Examples:
The analog output signal has a voltage which depends on gas concentration relative to the full scale of the range. To relate the signal output to
the actual concentration, it is necessary to know what range the instrument
is currently on, especially when the analyzer is in the autoranging mode.
The signal output for concentration is linear over the currently selected analysis range. For example, if the analyzer is set on a range that
was defined as 0–10 % hydrogen, then the output would be as shown in
Table 3-1.
Table 3-1: Analog Concentration Output—Example
Percent
Hydrogen
Voltage Signal
Output (V dc)
0
1
2
3
4
5
6
7
8
9
10
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
Current Signal
Output (mA dc)
4.0
5.6
7.2
8.8
10.4
12.0
13.6
15.2
16.8
18.4
20.0
Teledyne Analytical Instruments
3-5
3 Installation
Model 2020
To provide an indication of the range, the Range ID analog output
terminals are used. They generate a steady preset voltage (or current when
using the current outputs) to represent a particular range. Table 3-2 gives
the range ID output for each analysis range.
Table 3-2: Analog Range ID Output—Example
Range
Range 1
Voltage (V)
0.25
Current (mA) Application
8
0-1 % H2 in N2
Range 2
0.50
12
0-10 % H2 in N2
Range 3
0.75
16
0-1 % H2 in Air
Range 4 (Cal)
1.00
20
0-1 % H2 in N2
3.3.5 Alarm Relays
The three alarm-circuit connectors are spring terminals for making
connections to internal alarm relay contacts. Each provides a set of Form
C contacts for each type of alarm. Each has both normally open and normally closed contact connections. The contact connections are indicated by
diagrams on the rear panel. They are capable of switching up to 3 amperes
at 250 V ac into a resistive load. See Figure 3-5. The connectors are:
Threshold Alarm 1:
• Can be configured as high (actuates when concentration is above threshold), or low (actuates when
concentration is below threshold).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
Threshold Alarm 2:
• Can be configured as high (actuates when concentration is above threshold), or low (actuates when
concentration is below threshold).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
System Alarm:
Actuates when DC power supplied to circuits is
unacceptable in one or more parameters. Permanently configured as fail-safe and latching. Cannot
be defeated.
Actuates when cell can not balance during zero
calibration.
Actuates when span parameter out off its limited
parameter.
3-6
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Installation 3
Actuates when self test fails.
To reset a system alarm, call out the set up menue
by scroll keys. Use UP/DOWN key to select
STANDBY function. Turn off analyzer by pressing
ENTER key. Turn analyzer back on by selecting
any key. Set ESC key twice.
Further detail can be found in chapter 4, section 4-5.
DANGEROUS VOLTAGES MAY STILL BE
PRESENT AT THIS TERMINALS EVEN IF POWER
TO THE INSTRUMENT IS REMOVED.
Figure 3-5: Types of Relay Contacts
3.3.6 Digital Remote Cal Inputs
Accept 0 V (off) or 24 V dc (on) inputs for remote control of calibration. (See Remote Calibration Protocol below.)
Zero:
Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the Zero mode. Either side may be
grounded at the source of the signal. A synchronous signal
must open and close the external gas control valves appropriately. See 3.3.9 Remote Probe Connector. (With the –C
option, the internal valves operate automatically.)
Span:
Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the Span mode. Either side may be
grounded at the source of the signal. A synchronous signal
must open and close the external gas control valves appro-
Teledyne Analytical Instruments
3-7
3 Installation
Model 2020
priately. See 3.3.9 Remote Probe Connector. (With the –C
option, the internal valves operate automatically.)
Cal Contact: This relay contact is closed while analyzer is spanning
and/or zeroing. (See Remote Calibration Protocol below.)
Remote Calibration Protocol: To properly time the Digital Remote
Cal Inputs to the Model 2020 Analyzer, the customer's controller must
monitor the Cal Relay Contact.
When the contact is OPEN, the analyzer is analyzing, the Remote Cal
Inputs are being polled, and a zero or span command can be sent.
When the contact is CLOSED, the analyzer is already calibrating. It
will ignore your request to calibrate, and it will not remember that request.
Once a zero or span command is sent, and acknowledged (contact
closes), release it. If the command is continued until after the zero or span
is complete, the calibration will repeat and the Cal Relay Contact (CRC)
will close again.
When the contact is closed, the display would display the last reading
of the gas concentration value and output signal would output the last
reading from the sample gas (SAMPLE and HOLD).
For example:
1) Test the CRC. When the CRC is open, Send a zero command
until the CRC closes (The CRC will close quickly.)
2) When the CRC closes, remove the zero command.
3) When CRC opens again, send a span command until the CRC
closes. (The CRC will close quickly.)
4) When the CRC closes, remove the span command.
When CRC opens again, zero and span are done, and the sample is
being analyzed.
Note: The Remote Probe connector (paragraph 3.3.9) provides
signals to operate the zero and span gas valves synchronously. However, if you have the –C Internal valve option,
which includes zero and span gas inputs, the 2020 automatically regulates the zero, span and sample gas flow.
3.3.7 Range ID Relays
Four dedicated Range ID relay contacts. For any single application
they are assigned to relays in ascending order. For example: if all ranges
have the same application, then the lowest range is assigned to the Range 1
3-8
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Installation 3
ID relay, and the highest range is assigned to the Range 3 ID relay. Range
4 is the Cal Range ID relay.
3.3.8 Network I/O
A serial digital input/output for local network protocol. At this printing, this port is not yet functional. It is to be used in future versions of the
instrument.
3.3.9 RS-232 Port
The digital signal output is a standard RS-232 serial communications
port used to connect the analyzer to a computer, terminal, or other digital
device. Pin outs are listed in Table 3-3.
Table 3-3: RS-232 Signals
RS-232 Sig
DCD
RD
TD
DTR
COM
DSR
RTS
CTS
RI
RS-232 Pin
1
2
3
4
5
6
7
8
9
Purpose
Data Carrier Detect
Received Data
Transmitted Data
Data Terminal Ready
Common
Data Set Ready
Request to Send
Clear to Send
Ring Indicator
Output: The data output is status information, in digital form, updated every two seconds. Status is reported in the following order:
• The concentration in ppm or percent
• Type of gas
• The range in use (01 = Range 1, 02 = Range 2, 03 = Range 3,
CAL = Range 4)
• The scale of the range (0-100 %, etc)
• Which alarms—if any—are disabled (AL–x OFF)
• Which alarms—if any—are tripped (AL–x ON).
Each status output is followed by a carriage return and line feed.
Input: The input functions using RS-232 that have been implemented to date are described in Table 3-3.
Teledyne Analytical Instruments
3-9
3 Installation
Model 2020
Table 3-4: Commands via RS-232 Input
Command
Description
as
Immediately starts an autospan.
az
Immediately starts an autozero.
rp
Allows reprogramming of the APPLICATION (gas use) and
ALGORITHM (linearization) System functions.
st
Toggling input. Stops/Starts any status message output
from the RS-232, until st is sent again.
rm1
Range manual 1
rm2
Range manual 2
rm3
Range manual 3
rm4
Range manual CAL
ra
Range auto
Implementation: The RS-232 protocol allows some flexibility in its
implementation. Table 3-4 lists certain RS-232 values that are required by
the Model 2020 implementation.
Table 3-5: Required RS-232 Options
Parameter
Baud
Byte
Parity
Stop Bits
Message Interval
Setting
2400
8 bits
none
1
2 seconds
3.3.10 Remote Probe Connector
The 2020 is a single-chassis instrument, which has no Remote Probe
Unit. Instead, the Remote Probe connector is used as another method for
controlling external sample/zero/span gas valves. See Figure 3-6.
3-10
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Installation 3
Figure 3-6: Remote Probe Connector Pinouts
The voltage from these outputs is nominally 0 V for the OFF and
15 V dc for the ON conditions. The maximum combined current that can
be pulled from these output lines is 100 mA. (If two lines are ON at the
same time, each must be limited to 50 mA, etc.) If more current and/or a
different voltage is required, use a relay, power amplifier, or other matching circuitry to provide the actual driving current.
In addition, each individual line has a series FET with a nominal ON
resistance of 5 ohms (9 ohms worst case). This could limit the obtainable
voltage, depending on the load impedance applied. See Figure 3-7.
Figure 3-7: FET Series Resistance
3.4
Gas Connections
The gas fittings are accessed through holes on the underside of the
analyzer chassis, as shown in Figure 3-8. Use 1/8 NPT threaded conversion
fittings to convert pipe to tube for these connectors.
Teledyne Analytical Instruments
3-11
3 Installation
Model 2020
Figure 3-8: Gas Connections to the Basic Unit
There are no gas control valves inside the main chassis. A sample
system must be provided for introduction of zero and span gas, as well as
sample gas, into the sample path, and for controlling the flowrates through
the sample and reference paths of the analyzer.
If you have purchased a gas selector panel from Analytical Instruments, the drawings at the back of this manual will contain a dimension
drawing, with the modified cutout and hole pattern for mounting, and a
drawing and/or addendum showing the gas connections. Figure 3-9 is an
example showing a manual-valve panel with three valves (for sample, span
and zero gases) and two flowmeters (one for reference and one for sample,
span and zero gases).
3-12
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Installation 3
Figure 3-9: Front Panel with optional selector panel (as shown)
3.4.1 Sample System Design
Gas Connector and Selector Panels for specific applications are
available at additional cost . These panels are optional designed to substitute a standard front panel.
For those customers wishing to incorporate their own sample system,
electronic input/output ports are provided on the electrical connection
board for the operation of solenoid valves under the complete control of
the Model 2020 electronics. See section 3.3. The recommended system
piping schematic is included among the drawings at the rear of the manual.
The unit is manufactured with 1/4 inch tubing and 1/8 NPT thread ports.
The customer must provide matching fittings.
For best results, use the recommended piping system. Select a
flowmeter that can resolve 0.08 scfh (40-50 cc/min) for the reference path
of the analyzer, and select a flowmeter that can resolve 0.3 scfh (150 cc/
min) for the sample path of the analyzer.
Note: The sample-line pressure regulator should be installed as
close to the sample point as possible to minimize sample-line
lag time.
Teledyne Analytical Instruments
3-13
3 Installation
Model 2020
NOTE: An additional option is available for SEALED reference application. This option would not have the reference Gas Flow
Meter, Piping and Fittings.
3.4.2 Pressure and Flowrate Regulation
Appropriate pressure reducing regulators must be installed at all gas
supply sources. To minimize flowrate adjustments the pressure regulators
on the supporting gas supply cylinders should be adjusted to provide the
same output pressure as the sample line regulator.
The gas pressure in should be reasonably well regulated. Pressures
between 5 and 50 psig are acceptable (10 psig is normal) as long as the
pressure, once established, will keep the flow constant during analysis
and within an acceptable range (between 0.1 and 0.4 scfh—See Note).
Note: Gases lighter than air have a flowrate higher than indicated
on the flowmeter, while gases heavier than air have a flowrate
lower than indicated. Values can range from one half to twice
the indicated flowrate.
For example: For hydrogen or helium, set the flowrate to 0.1
scfh (50 cc/min). For carbon dioxide or argon, set the flowrate
to 0.4 scfh (200 cc/min).
When installing pressure regulators on supply cylinders, crack the
cylinder valves so that gas is flowing during installation. This will eliminate the most common cause of standardization-gas contamination: air
trapped during assembly diffusing back into the cylinder. This procedure is
particularly important in applications where impurity content of 1 to 2 % is
the range of interest.
Note: If you have the –V option, The above pressure and flow
values apply instead to the vacuum at the VENT connector,
described below, with minus signs before the pressure
readings.
3.4.3 VENT Exhaust
There are two separate VENT fittings—one for the sample gas and
one for the reference gas. Use 1/4 inch tubing for both sample and reference
vents to minimize back pressure from restricted flow.
Exhaust connections must be consistent with the hazard level of the
constituent gases. Check local, state, and federal laws, and ensure that the
exhaust stream vents to an appropriately controlled area if required. If not
vented to the same area, both VENT lines must vent to areas with equal
3-14
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Installation 3
ambient pressures, and pressures must vary no more than the normal
barometric changes.
Install VENT lines such that water and dirt cannot accumulate in
them.
Note: If your 2020 has the –V option, see Note at end of Pressure
and Flow Rate Regulation, above, for gas vacuum/flow
considerations.
3.4.4 SAMPLE Gas
In the standard model, sample and calibration gases are introduced
through the SAMPLE fitting. The gases must be Tee'd into the Sample inlet
with appropriate valves.
The gas pressure in should be well regulated. (See section 3.4.1.) The
sample line pressure regulator should be installed as close to the sample
line as possible to minimize sample line lag time.
If greater flow is required for improved response time, install a bypass
in the sampling system upstream of the analyzer input.
3.4.5 REFERENCE Gas
A gas of fixed composition is needed as a reference to which the
sample gas will be compared. The reference gas is normally selected to
represent the main background gas of the analysis.
For most applications, a constant supply of reference gas flowing at
the same rate as the sample is required for best results. However, in many
cases the flow of reference gas can be slowed to about 0.08 scfh
(40 cc/min) with good results.
For some applications, an optional sealed air reference is installed. In
sealed-reference sensors the reference side of the detector cell is filled with
air and sealed. This eliminates the need to have reference gas constantly
passing through the cell.
NOTE: For instruments equipped with the optional sealed air reference, there is no REFERENCE inlet or reference VENT port.
It is highly recommended that the same cylinder of gas be used for
both the REFERENCE gas and the ZERO gas.
Pressure, flow, and safety considerations are the same as prescribed
for the SAMPLE gas, above.
Teledyne Analytical Instruments
3-15
3 Installation
Model 2020
3.4.6 ZERO Gas
For the ZERO gas, a supply of the background gas, usually containing
none of the impurity, is required to zero the analyzer during calibration.
For suppressed zero ranges the zero gas must contain the low-end concentration of the impurity.
NOTE: Because most cylinder gases are between 99.95 and 99.98%
pure, it is highly recommended that the same cylinder of gas
be used for both REFERENCE and ZERO gas.
NOTE: It is essential to the accuracy of the analyzer that the purity of
the zero gas be known. Otherwise, when the zero control is
adjusted during zero standardization, the reading will indicate
the impurity content of the zero gas, rather than zero.
3.4.7 SPAN Gas
For the SPAN gas, a supply of the background gas containing 70100 % of the component of interest is required as a minimum.
Note: If your analyzer range is set for inverting output, your zero
gas will be at 100% of the range interest, and span will be 70
to 100% of the low end range.
If linearization is required, intermediate concentrations of the target
gas in the background gas may be necessary. From one to nine separate
span gases may be used, depending on the desired precision of the linearization. See chapter 4, Operation.
3.5
Testing the System
Before plugging the instrument into the power source:
• Check the integrity and accuracy of the gas connections. Make
sure there are no leaks.
• Check the integrity and accuracy of the electrical connections.
Make sure there are no exposed conductors
• Check that the pressure and flow of all gases are within the
recommended levels, and appropriate for your application.
Power up the system, and test it by performing the following
operations:
1. Repeat the Self-Diagnostic Test as described in chapter 4,
section 4.3.5.
3-16
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Operation 4
Operation
4.1
Introduction
Although the Model 2020 is usually programmed to your application at
the factory, it can be further configured at the operator level, or even,
cautiously, reprogrammed. Depending on the specifics of the application,
this might include all or a subset of the following procedures:
• Setting system parameters:
• Establish a security password, if desired, requiring Operator
to log in.
• Establish and start an automatic calibration cycle, if desired.
• Routine Operation:
• Calibrate the instrument.
• Choose autoranging or select a fixed range of analysis.
• Set alarm setpoints, and modes of alarm operation (latching,
fail-safe, etc).
• Program/Reprogram the analyzer:
• Define new applications.
• Linearize your ranges.
• Special functions setup:
• Set output reversal.
• Set polarity reversal or offset output.S
• Set gain amplification.
Before you configure your 2020, the following default values are in
effect: RANGE/APPLICATIONS: refer to data sheet on the first page of
this manual;
Range:
Manual
Alarm Relays:
Defeated, 0.00%, HI, NOT Fail/Safe, not latching
Zero:
Auto, every 0 days 0 hours
Teledyne Analytical Instruments
4-1
4 Operation
Span:
Model 2020
Auto, at 10%, every 0 days, at 0 hours
Password: TAI
4.2
Using the Controls
To get the proper response from these controls, turn the control
toward the desired action (ESCAPE or ENTER—DOWN or UP), and then
release it. Turn-and-release once for each action. For example, turn-andrelease twice toward UP to move the VFD screen two selections upwards
on the list of options (menu).
The item that is between arrows on the screen is the item that is currently selectable by choosing ENTER (turn-and-release toward ENTER
with the ESCAPE/ENTER control).
In these instructions, to ENTER means to turn-and-release toward
ENTER, and To ESCAPE means to turn-and-release towards ESCAPE. To
scroll UP (or scroll DOWN) means to turn-and-release toward UP (or
DOWN) as many times as necessary to reach the required menu item.
4.2.1 Mode/Function Selection
When the analyzer is first powered up, and has completed its initialization and self diagnostics, ESCAPE toggles the instrument between the
ANALYZE screen (Analysis Mode) and the MAIN MENU screen (Setup
Mode). The ANALYZE screen is the only screen of the Analysis Mode.
The MAIN MENU screen is the top level in a series of screens used in
the Setup Mode to configure the analyzer for the specific application. The
DOWN/UP commands scroll through the options displayed on the VFD
screen. The selectable option appears between arrows. When you reach the
desired option by scrolling, ENTER the selection as described below.
ESCAPE takes you back up the hierarchy of screens until you reach
the ANALYZE MODE. ESCAPING any further just toggles between the
MAIN MENU and the ANALYZE screen.
4.2.1.1 Analysis Mode
This is the normal operating mode. The analyzer monitors the oconcentration of the mixure content of the sample, displays the percent of the
concentration in the sample stream, and warns of any alarm conditions.
4-2
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Operation 4
SETUP MODE
Span/Zero
Off/On
Span/Zero
Timing
PSWD
Enter
Password
Change
Yes/No
LOGOUT
Secure Sys &
AnalyzeOnly
MODEL
Show Model
andVersion
SELF-TEST
Self-Test in
Progress
AUTO-CAL
Auto/Manual
SpanSelect
SpanValue
Set
ZERO
Auto/Manual
ZeroSelect
Zeroin
Progress
ALARMS
Select
Range
Gas Use
Range
RANGE
APPLICATION
ALOGORITHM
Change
Password
Verify
Password
Spanin
Progress
%/ppm
Select
Setpoints&
Attributes
Define
Range
Select
Range
Auto/Manual
RangeAdj
Auto
CONTRAST
Yes
Slef-Test
Results
SPAN
Man
Span/Zero
Off/On
Set LCD
Contrast
Gas
Application
Contrast Function is DISABLED
(Refer to Section 1.6)
Select
Range
Define
Appl/Range
Gas Use
Range
Select
Range
Select
OFF/INV
Ver
Select
Verify/Setup
Verify
Points
Enter
Man Input/Output
Values
Enter
Auto/Manual
Set LinearityCal
CAL-INDPD
STAND-BY
Calibrateone
rangeatatime
SelectLinrty
Auto SpanValues
Enter
ONw/out
displays/outputs
Figure 4-1: Hierarchy of Functions and Subfunctions
Teledyne Analytical Instruments
4-3
4 Operation
Model 2020
Either control switches you to Setup Mode. Setup Mode switches back to
Analyze Mode if no controls are used for more than five seconds.
4.2.1.2 Setup Mode
The MAIN MENU consists of 14 functions you can use to customize
and check the operation of the analyzer. Figure 4-1 shows the functions
available with the 2020. They are listed here with brief descriptions:
1 AUTO-CAL: Used to define and/or start an automatic
calibration sequence.
2 PSWD: Used to establish password protection or change the
existing password.
3 LOGOUT: Logging out prevents unauthorized tampering with
the analyzer settings.
4 MODEL: Displays Manufacturer, Model, and Software version
of the instrument.
5 SELF-TEST: The instrument performs a self-diagnostic
routine to check the integrity of the power supply, output
boards, cell and amplifiers.
6 SPAN: Set up and/or start a span calibration.
7 ZERO: start a zero calibration.
8 ALARMS: Used to set the alarm setpoints and determine
whether each alarm will be active or defeated, HI or LO acting,
latching or not, and failsafe or not.
9 RANGE: Used to set up three analysis ranges that can be
switched automatically with auto-ranging or used as individual
fixed ranges.
10 CONTRAST: Increase or decrease the LCD screen contrast.
YOU MAY NEED TO DO THIS AT TURN-ON. See Setting
the Display Contrast, below.
11 APPLICATIONS: Restricted function, not generally accessed
by the end user. Used to define up to three analysis ranges and
a calibration range (including impurity, background low end of
range, high end of range, and % of ppm units).
12 ALOGORITHM: Arestricted function, not generally accessed
by the end user. Used to linearize the output for the range of
interest.
4-4
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Operation 4
13 CAL-INDEPD: Not generally accessed buy the end user.
Forces analyzer to be in independent calibration mode.
14 STANDBY: Remove power to outputs and displays, but
maintain power to internal circuitry.
Any function can be selected at any time. Just scroll through the MAIN
MENU with the DOWN/UP control to the appropriate function, and
ENTER it. The analyzer will immediately start that function, unless
password restrictions have been assigned. (Password assignment is
explained further on.)
All of these functions are described in greater detail in the procedures
starting in section 4.3. The VFD screen texts used to illustrate the
procedures are reproduced in a Monospaced type style.
4.2.2 Data Entry
4.2.2.1 ENTER
When the selected option is a function on the Main Menu screen, the
function name appears between the arrows on the screen. You activate the
function by turning the ESCAPE/ENTER control to ENTER.
When the selected option is a function or subfunction, ENTER
moves the display to the VFD screen for that function or subfunction.
When the selected option is a modifiable item, the DOWN/UP control
can be used to increment or decrement that modifiable item to the value or
action you want. Then you ENTER the item, which also puts you into the
next field to continue programming.
When the last field is entered, ENTER takes you to the next screen in
the process, or if the process is completed, ENTER takes you back to the
ANALYZE screen.
4.2.2.2 ESCAPE
A turn-and-release toward ESCAPE moves the blinking to the next
field on the left. When you are on the leftmost field, another ESCAPE takes
you back to the previous screen.
If you do not wish to continue a function, you can abort the session by
escaping to the leftmost field, and then issuing another ESCAPE. Escaping
Teledyne Analytical Instruments
4-5
4 Operation
Model 2020
a function takes the analyzer back to the previous screen, or to the ANALYZE Function, depending on the function escaped.
reproduced, at the appropriate point in the procedure, in a Monospaced type style. Push-button names are printed in Oblique type.
4.3.1 Setting the Display
Contrast Function is DISABLED
(Refer to Section 1.6)
If you cannot read anything on the display after first powering up:
1. Observe LED readout.
a. If LED meter reads all eights and dots, go to step 3.
b. If LED meter displays anything else, go to step 2.
2. Disconnect power to the Analyzer and reconnect again. LED
meter should now read all eights and dots.
4.3.2 Setting up an AUTO-CAL
When proper automatic valving is connected (see chapter 3, installation), the Analyzer can cycle itself through a sequence of steps that automatically zero and span the instrument.
Note: Before setting up an AUTO-CAL, be sure you understand the
Zero and Span functions as described in section 4.4, and
follow the precautions given there.
Note: If you require highly accurate AUTO-CAL timing, use external
AUTO-CAL control where possible. The internal clock in the
Model 2020 is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.
Note: If your ranges are configured for different applications, then
AUTO-CAL will calibrate all of the ranges simultaneously (by
calibrating the Cal Range).
To setup an AUTO-CAL cycle:
The VFD will display five subfunctions.
4-6
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Operation 4
Call out MAIN MENU, scroll to AUTO-CAL function, and ENTER.
A new screen for ZERO/SPAN set appears.
ZERO in
SPAN in
Ød
Ød
Øh off
Øh off
Use UP/DOWN Control to blink ZERO (or SPAN), then Enter. (You
won’t be able to set OFF to ON if a zero interval is entered.) A Span Every
... (or Zero Every ...) screen appears.
Zero schedule: OFF
Day:
Ød Hour:
Øh
Use UP/DOWN Control to set a value in days, then ENTER to move
to the start-time value in hours. Use UP/DOWN to set a start-time value,
then ENTER.
To turn ON the SPAN and/or ZERO cycles (to activate AUTOCAL):
useUP/DOWN Control to set the OFF/ON field to ON. You can now turn
these fields ON because there is a nonzero span time defined.
4.3.3 Password Protection
Before a unique password is assigned, the system assigns TAI by
default. This password will be displayed automatically. The operator just
presses the Enter key to be allowed total access to the instrument’s features.
If a password is assigned, then setting the following system parameters
can be done only after the password is entered: alarm setpoints, AUTOCAL setup. ZERO/SPAN calibration assigning a new password, range/
application selections, and curve algorithm linearization. (APPLICATION and ALGORITHM are covered in the programming section.) However, the instrument can still be used for analysis or for initiating a self-test
without entering the password. To defeat security the password must be
changed back to TAI.
NOTE: If you use password security, it is advisable to keep a copy of
the password in a separate, safe location.
4.3.3.1
Entering the Password
To install a new password or change a previously installed password,
you must key in and ENTER the old password first. If the default password
is in effect, pressing the ENTER button will enter the default TAI password
for you.
Call out MAIN MENU setup by selecting any controls
Teledyne Analytical Instruments
4-7
4 Operation
Model 2020
Use the UP/DOWN key to scroll the blinking over to PSWD, and press
Enter to select the password function. Either the default TAI password or
AAA place holders for an existing password will appear on screen depending
on whether or not a password has been previously installed.
Enter password:
T A I
or
Enter password:
A A A
The screen prompts you to enter the current password. If you are not
using password protection, press Enter to accept TAI as the default password. If a password has been previously installed, enter the password using
the ENTER key to scroll through the letters, and the UP/DOWN key to
change the letters to the proper password. The last ENTER enters the
password.
In a few seconds, you will be given the opportunity to change this
password or keep it and go on.
Change Password?
=Yes
=No
Press Escape to move on, or proceed as in Changing the Password,
below.
4.3.3.2
Installing or Changing the Password
If you want to install a password, or change an existing password,
proceed as above in Entering the Password. When you are given the opportunity to change the password:
Change Password?
=Yes
=No
Enter to change the password (either the default TAI or the previously
assigned password), or press Escape to keep the existing password and
move on.
If you chose Enter to change the password, the password assignment
screen appears.
Select new password
T A I
or
4-8
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Operation 4
Select new password
AAA
Enter the password using theUP/DOWN and ENTER to scroll
through the existing password letters, and the UP/DOWN keys to change
the letters to the new password. The full set of 94 characters available for
password use are shown in the table below.
Characters Available for Password Definition:
A
K
U
_
i
s
}
)
3
=
B
L
V
`
j
t
→
*
4
>
C
M
W
a
k
u
!
+
5
?
D
N
X
b
l
v
"
'
6
@
E
O
Y
c
m
w
#
7
F
P
Z
d
n
x
$
.
8
G
Q
[
e
o
y
%
/
9
H
R
¥
f
p
z
&
0
:
I
S
]
g
q
{
'
1
;
J
T
^
h
r
|
(
2
<
When you have finished typing the new password, press Enter. A
verification screen appears. The screen will prompt you to retype your
password for verification.
Enter PWD To Verify:
A A A
Use the UP/DOWN key to retype your password and use ENTER to
scroll through the letters, and last enter will complete verification. Your
password will be stored in the microprocessor and the system will immediately switch to the Analyze screen, and you now have access to all instrument functions.
If all alarms are defeated, the Analyze screen appears as:
1.95
nR1:
% H2 in N2
Ø 1Ø Anlz
If an alarm is tripped, the second line will change to show which alarm
it is:
1.95
AL1
% H2 in N2
NOTE: If you log off the system using the LOGOUT function in the
MAIN MENU, you will now be required to reenter the password
to gain access to Alarm, and Range functions.
Teledyne Analytical Instruments
4-9
4 Operation
Model 2020
4.3.4 Logging Out
The LOGOUT function provides a convenient means of leaving the
analyzer in a password protected mode without having to shut the instrument off. By entering LOGOUT, you effectively log off the instrument
leaving the system protected against use until the password is reentered. To
log out, scroll to field of LOGOUT function, and ENTER to logout The
screen will display the message:
Protected until
password entered
4.3.5 System Self-Diagnostic Test
The Model 2020 has a built-in self-diagnostic testing routine. Preprogramming signals are sent through the power supply, output board, preamp
board and sensor circuit. The return signal is analyzed, and at the end of the
test the status of each function is displayed on the screen, either as OK or as
a number between 1 and 1024. (See System Self Diagnostic Test in chapter
5 for number code.) If any of the functions fails, the System Alarm is
tripped.
Note: The sensor will always show failed unless identical gas is
present in both channels at the time of the SELF-TEST.
The self diagnostics are run automatically by the analyzer whenever the
instrument is turned on, but the test can also be run by the operator at will.
To initiate a self diagnostic test during operation, use the UP/DOWN key to
scroll through the MAIN MENU to the SELFTEST and Enter. The screen
will follow the running of the diagnostic.
RUNNING DIAGNOSTIC
Testing Preamp Cell
When the testing is complete, the results are displayed.
Power: OK Analog: OK
Cell:
2 Preamp: 3
The module is functioning properly if it is followed by OK. A number
indicates a problem in a specific area of the instrument. Refer to Chapter 5
Maintenance and Troubleshooting for number-code information. The
results screen alternates for a time with:
Press Any Key
To Continue...
4-10
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Operation 4
Then the analyzer returns to the initial System screen.
4.3.6 The Model Screen
Scroll through the MAIN MENU to MODEL and Enter. The screen
displays the manufacturer, model, and software version information.
4.3.7 Checking Linearity with ALGORITHM
Use UP/DOWN control to select ALGORITHM, and Enter.
sel rng to set algo:
> Ø1 Ø2 Ø3
<
Use the UP/DOWN Control to select the range: 01, 02, or 03. Then
press Enter.
Gas Use:
H2 N2
Range:
Ø 10%
Enter again.
Algorithm setup:
VERIFY
SET UP
Use UP/DOWN key to select and Enter VERIFY to check whether the
linearization has been accomplished satisfactorily.
Dpt
Ø
INPUT
Ø.ØØ
OUTPUT
Ø.ØØ
The leftmost digit (under Dpt) is the number of the data point being
monitored. Use the UP/DOWN key to select the successive points.
The INPUT value is the input to the linearizer. It is the simulated output
of the analyzer. You do not need to actually flow gas.
The OUTPUT value is the output of the linearizer. It should be the
ACTUAL concentration of the span gas being simulated.
If the OUTPUT value shown is not correct, the linearization must be
corrected. ESCAPE to return to the previous screen. Select and Enter SET
UP to Calibration Mode screen.
Select algorithm
mode : AUTO
There are two ways to linearize: AUTO and MANUAL: The auto mode
requires as many calibration gases as there will be correction points along
Teledyne Analytical Instruments
4-11
4 Operation
Model 2020
the curve. The user decides on the number of points, based on the precision
required.
The manual mode only requires entering the values for each correction
point into the microprocessor via the front panel buttons. Again, the number
of points required is determined by the user.
4.4
The Zero and Span Functions
(1) The Model 2020 can have as many as three analysis ranges plus a
special calibration range (Cal Range); and the analysis ranges, if more than
one, may be programmed for separate or identical gas applications.
(2) If all ranges are for the same application, then you will not need the
Cal Range. Calibrating any one of the ranges will automatically calibrate the
others.
(3) If: a) each range is programmed for a different gas application, b)
your sensor calibration has drifted less than 10 %, and c) your Cal Range
was calibrated along with your other ranges when last calibrated, then you
can use the Cal Range to calibrate all applications ranges at once.
If your Model 2020 analyzer fits the paragraph (3) description, above,
use the Cal Range. If your analyzer has drifted more than 10 %, calibrate
each range individually.
CAUTION: Always allow 4-5 hours warm-up time before calibrating, if your analyzer has been disconnected
from its power source. This does not apply if the
analyzer was plugged in but was in STANDBY.
The analyzer is calibrated using reference, zero, and span gases. Gas
requirements are covered in detail in chapter 3, section 3.4 Gas
Connections. Check that calibration gases are connected to the analyzer
according to the instructions in section 3.4, observing all the prescribed
precautions.
Note: Shut off the gas pressure before connecting it to the analyzer,
and be sure to limit pressure to 40 psig or less when turning it
back on.
Readjust the gas pressure into the analyzer until the flowrate through
the sensor settles between 50 to 200 cc/min (approximately 0.1 to 0.4 scfh).
Note: Always keep the zero calibration gases flow as close as the
flowrate of sample gas as possible
4-12
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Operation 4
4.4.1 Zero Cal
The ZERO function in the MAIN MENU is used to enter the zero
calibration function. Zero calibration can be performed in either the automatic or manual mode.
CAUTION: If you are zeroing the Cal Range by itself (multiple
application analyzers only), use manual mode
zeroing.
If you want to calibrate ALL of the ranges at once
(multiple application analyzers only), use auto mode
zeroing in the Cal Range.
Make sure the zero gas is flowing to the instrument. If you get a CELL
CANNOT BE BALANCED message while zeroing skip to section 4.4.1.3.
4.4.1.1
Auto Mode Zeroing
Observe the precautions in sections 4.4 and 4.4.1, above.Scroll to
ZERO function buy using UP/DOWN control and enter the zero function
mode. The screen allows you to select whether the zero calibration is to be
performed automatically or manually. Use the UP/DOWN key to toggle
between AUTO and MAN zero settling. Stop when AUTO appears, blinking,
on the display.
Select zero
mode: AUTO
Press Enter to begin zeroing.
####.## %
H2 N2
Slope=#.###
CZero
The beginning zero level is shown in the upper left corner of the
display. As the zero reading settles, the screen displays and updates information on Slope= in ppm/second (unless the Slope starts within the acceptable zero range and does not need to settle further). The system first does a
course zero, shown in the lower right corner of the screen as CZero, for
approximate 3 min, and then does a fine zero, and displays FZero, for
approximate 3 min.
Then, and whenever Slope is less than 0.01 for at least 3 min, instead
of Slope you will see a countdown: 9 Left, 8 Left, and so fourth. These are
software steps in the zeroing process that the system must complete, AF-
Teledyne Analytical Instruments
4-13
4 Operation
Model 2020
TER settling, before it can go back to Analyze. Software zero is indicated
by SZero in the lower right corner.
####.## %
H2 N2
4 Left=#.### SZero
The zeroing process will automatically conclude when the output is
within the acceptable range for a good zero. Then the analyzer automatically returns to the Analyze mode.
4.4.1.2
Manual Mode Zeroing
Scroll to Zero and enter the Zero function. The screen that appears
allows you to select between automatic or manual zero calibration. Use the
UP/DOWN keys to toggle between AUTO and MAN zero settling. Stop when
MANUAL appears, blinking, on the display.
Select zero
mode: MANUAL
Enter to begin the zero calibration. After a few seconds the first of
three zeroing screens appears. The number in the upper left hand corner is
the first-stage zero offset. The microprocessor samples the output at a
predetermined rate.
####.##
% H2 N2
Zero adj:2048 CZero
The analyzer goes through C–Zero, F–Zero, and S–Zero. During C–
Zero and F–Zero, use the UP/DOWN keys to adjust displayed Zero adj:
value as close as possible to zero. Then, press Enter.
S–Zero starts. During S–Zero, the Microcontroller takes control as in
Auto Mode Zeroing, above. It calculates the differences between successive
samplings and displays the rate of change as Slope= a value in parts per
million per second (ppm/s).
####.##
%
Slope=#.###
H2 N2
SZero
Generally, you have a good zero when Slope is less than 0.05 ppm/s for
about 30 seconds.
Once zero settling completes, the information is stored in the analyzer’s memory, and the instrument automatically returns to the Analyze
mode.
4-14
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Operation 4
4.4.1.3 Cell Failure
Cell failure in the 2020 is usually associated with inability to zero the
instrument with a reasonable voltage differential across the Wheatstone
bridge. If this should ever happen, the 2020 system alarm trips, and the
VFD displays a failure message.
Cell cannot be balanced
Check your zero gas
Before replacing the sensor:
a. Check your zero gas to make sure it is within specifications.
b. Check for leaks downstream from the sensor, where contamination may be leaking into the system.
c. Check flowmeter to ensure that the flow is no more than
200SCCM
d. Check temperature controller board.
e. Check gas temperature.
If none of the above as indicated, the sensor may need to be replaced.
Check warranty, and contact Analytical Instruments Customer Service.
4.4.2 Span Cal
The Span button on the front panel is used to span calibrate the
analyzer. Span calibration can be performed in either the automatic or
manual mode.
CAUTION: If you are spanning the Cal Range by itself (multiple
application analyzers only), use manual mode
zeroing.
If you want to calibrate ALL of the ranges at once
(multiple application analyzers only), use auto mode
spanning in the Cal Range.
Make sure the span gas is flowing to the instrument.
4.4.2.1
Auto Mode Spanning
Observe all precautions in sections 4.4 and 4.4.2, above. Scroll SPAN
and enter the span function. The screen that appears allows you to select
whether the span calibration is to be performed automatically or manually.
Teledyne Analytical Instruments
4-15
4 Operation
Model 2020
Use the UP/DOWN key to toggle between AUTO and MAN span settling.
Stop when AUTO appears, blinking, on the display.
Select span
mode: AUTO
Enter to move to the next screen.
Span Val:
2Ø.ØØ %
To begin span
Use UP/DOWN key to change the span setting value.
ENTER will move the blinking field to units (%/ppm). Use UP/DOWN
key to select the units, as necessary. When you have set the concentration of
the span gas you are using, Enter to begin the Span calibration.
####.##%
Slope=#.###
H2
N2
Span
The beginning span value is shown in the upper left corner of the
display. As the span reading settles, the screen displays and updates information on Slope. Spanning automatically ends when the span output corresponds, within tolerance, to the value of the span gas concentration. Then
the instrument automatically returns to the analyze mode.
4.4.2.2
Manual Mode Spanning
Scroll Span by using UP/DOWN key and Enter to start the Span
function. The screen that appears allows you to select whether the span
calibration is to be performed automatically or manually.
Select span
mode: MANUAL
Use the UP/DOWN key to toggle between AUTO and MAN span
setting. Stop when MAN appears, blinking, on the display. ENTER to move
to next subfunction screen
Span Val: 2Ø.ØØ %
To begin span
Using the UP/DOWN key changes the span value, as necessary. Enter
to move to the units field (%/ppm). Use UP/DOWN key to select unit.
Press Enter to enter the span value into the system and begin the span
calibration.
Once the span has begun, the microprocessor samples the output at a
predetermined rate. It calculates the difference between successive sam-
4-16
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Operation 4
plings and displays this difference as Slope on the screen. It takes several
seconds for the first Slope value to display. Slope indicates rate of change of
the Span reading. It is a sensitive indicator of stability.
####.##%
H2 N2
Slope=#.###
Span
When the Span value displayed on the screen is sufficiently stable,
press Enter. (Generally, when the Span reading changes by 1 % or less of
the range being calibrated for a period of ten minutes it is sufficiently stable.) Once Enter is pressed, the Span reading changes to the correct value.
The instrument then automatically enters the Analyze function.
4.5
The Alarms Function
The Model 2020 is equipped with 6 fully adjustable set points concentration with two alarms and a system failure alarm relay. Each alarm relay
has a set of form “C" contacts rated for 3 amperes resistive load at 250
V ac. See Figure in Chapter 3, Installation and/or the Interconnection
Diagram included at the back of this manual for relay terminal connections.
The system failure alarm has a fixed configuration described in chapter
3 Installation.
The concentration alarms can be configured from the front panel as
either high or low alarms by the operator. The alarm modes can be set as
latching or non-latching, and either fail-safe or non-fail-safe, or, they can
be defeated altogether. The setpoints for the alarms are also established
using this function.
Decide how your alarms should be configured. The choice will depend
upon your process. Consider the following four points:
1. Which if any of the alarms are to be high alarms and which if
any are to be low alarms?
Setting an alarm as HIGH triggers the alarm when the
contaminant concentration rises above the setpoint. Setting an
alarm as LOW triggers the alarm when the contaminant
concentration falls below the setpoint.
Decide whether you want the alarms to be set as:
• Both high (high and high-high) alarms, or
• One high and one low alarm, or
• Both low (low and low-low) alarms.
2. Are either or both of the alarms to be configured as failsafe?
Teledyne Analytical Instruments
4-17
4 Operation
Model 2020
In failsafe mode, the alarm relay de-energizes in an alarm
condition. For non-failsafe operation, the relay is energized in
an alarm condition. You can set either or both of the
concentration alarms to operate in failsafe or non-failsafe mode.
3. Are either of the alarms to be latching?
In latching mode, once the alarm or alarms trigger, they will
remain in the alarm mode even if process conditions revert back
to non-alarm conditions. This mode requires an alarm to be
recognized before it can be reset. In the non-latching mode, the
alarm status will terminate when process conditions revert to
non-alarm conditions.
4. Are either of the alarms to be defeated?
The defeat alarm mode is incorporated into the alarm circuit so
that maintenance can be performed under conditions which
would normally activate the alarms.
The defeat function can also be used to reset a latched alarm.
(See procedures, below.)
If you are using password protection, you will need to enter your
password to access the alarm functions. Follow the instructions in section
4.3.3 to enter your password. Once you have clearance to proceed, enter
the Alarm function.
Note: If all ranges are for the same application, set any one of them
will automatically set the others.
Press the Alarm button on the front panel to enter the Alarm function.
Make sure that 01 is blinking.
Sel rng to set alm:
> Ø1 Ø2 Ø3
<
Set up the Range 1 alarm by moving the blinking over to 01 using the
UP/DOWN arrow keys. Then Enter. Check the gas application and range
limits as displayed on the screen.
Gas use: H2 N2
Range:
0 10 %
Press enter again to set the alarm setpoints.
Sel %/ppm alm to set
AL1PPM AL2PPM
Use the UP/DOWN keys to choose between % or ppm units. Then
Enter to move to the next screen.
4-18
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Operation 4
AL1: 1ØØØ
ppm HI
Dft:N Fs:N Ltch:N
Five parameters can be changed on this screen:
• Value of the alarm setpoint, AL1: ####
• Out-of-range direction, HI or LO
• Defeated? Dft:Y/N (Yes/No)
• Failsafe? Fs:Y/N (Yes/No)
• Latching? Ltch:Y/N (Yes/No).
• To define the setpoint, use the UP/DOWN key while screen is
blinking over to AL1: ####. Use the UP/DOWN key to change
the number. Holding down the key speeds up the incrementing
or decrementing.
• After the number (value) has been choosed, use Enter to move
the desired parameter. Then use the UP/DOWN keys to change
the parameter.
• Once the parameters for alarm have been set, Enter the alarm
function again, and repeat this procedure for next alarm.
• To reset a latched alarm, go to Dft and then use either UP two
times or DOWN two times. (Toggle it to Y and then back to N.)
–OR –
Go to Ltch and then use either UP two times or DOWN two
times. (Toggle it to N and back to Y.)
4.6
The Range Select Function
The Range function allows you to manually select the concentration
range of analysis (MANUAL), or to select automatic range switching (AUTO).
In the MANUAL screen, you are further allowed to define the high and
low (concentration) limits of each Range, and select a single, fixed range to
run.
CAUTION: If this is a linearized application, the new range
must be within the limits previously programmed
using the System function, if linearization is to
apply throughout the range. Furthermore, if the
limits are too small a part (approx 10 % or less) of
the originally linearized range, the linearization will
be compromised.
Teledyne Analytical Instruments
4-19
4 Operation
Model 2020
In the AUTO screen, you are further allowed to select which gas application (PREVIOUSLY defined in APPLICATION function) to run.
4.6.1 Manual (Select/Define Range) Screen
The Manual range-switching mode allows you to select a single, fixed
analysis range. It then allows you to redefine the upper and lower limits, for
the range.
Use UP/DOWN key to start the RANGE function, and ENTER
Select range
mode: MANUAL
Note: If all three ranges are currently defined for different application gases, then the above screen does not display (because
mode must be manual). Instead, the VFD goes directly to the
following screen.
If above screen displays, use the UP/DOWN arrow keys to Select
MANUAL, and press Enter.
Select range to run
> Ø1 Ø2 Ø3 CAL<
Use the UP/DOWN keys to select the range: 01, 02, 03, or CAL.
Then press Enter.
Gas use:
H2 N2
Range:
Ø 10 %
Use the ENTER key to move the range to low-end field. Use ENTER
key to move the range to high-end field. Use the UP/DOWN keys to
change the values of the fields.
Press Escape to return to the previous screen to select or define
another range.
Press Enter to return the to the Analyze function.
4.6.2 Auto (Single Application) Screen
Autoranging will automatically set to the application that has at least
two range setup with the same gases.
In the autoranging mode, the microprocessor automatically responds
to concentration changes by switching ranges for optimum readout sensitivity. If the upper limit of the operating range is reached, the instrument
4-20
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Operation 4
automatically shifts to the next higher range. If the concentration falls to
below 85% of full scale of the next lower range, the instrument switches to
the lower range. A corresponding shift in the DC concentration output, and
in the range ID outputs, will be noticed.
The autoranging feature can be overridden so that analog output stays
on a fixed range regardless of the contaminant concentration detected. If the
concentration exceeds the upper limit of the range, the DC output will
saturate at 1 V dc (20 mA at the current output).
However, the digital readout and the RS-232 output of the concentration are unaffected by the fixed range. They continue to read beyond the
full-scale setting until amplifier saturation is reached. Below amplifier
saturation, the over range readings are accurate UNLESS the application
uses linearization over the selected range.
The concentration ranges can be redefined using the Range function
Manual screen, and the application gases can be redefined using the APPLICATION function, if they are not already defined as necessary.
CAUTION: Redefining applications or ranges might require relinearization and/or recalibration.
To setup automatic ranging:
Select range on the MAIN MENU, and ENTER to start the Range
function.
Select range
mode : AUTO
Note: If all three ranges are currently defined for different application gases, then the above screen does not display (because
mode must be manual).
If above screen displays, use the UP/DOWN key to Select AUTO, and
Enter.
Press Escape to return to the previous Analyze Function.
Teledyne Analytical Instruments
4-21
4 Operation
Model 2020
4.6.3 Precautions
The Model 2020 allows a great deal of flexibility in choosing ranges
for automatic range switching. However, there are some pitfalls that are to
be avoided.
Ranges that work well together are:
• Ranges that have the same lower limits but upper limits that
differ by approximately an order of magnitude
• Ranges whose upper limits coincide with the lower limits of the
next higher range
• Ranges where there is a gap between the upper limit of the
range and the lower limit of the next higher range.
Range schemes that are to be avoided include:
• Ranges that overlap
• Ranges whose limits are entirely within the span of an adjoining
range.
Figure 4-2 illustrates these schemes graphically.
Figure 4-2: Examples of Autoranging Schemes
4.7
The Analyze Function
Normally, all of the functions automatically switch back to the Analyze function when they have completed their assigned operations. Pressing
4-22
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Operation 4
the Escape button in many cases also switches the analyzer back to the
Analyze function. Alternatively, if you leave your analyzer on MAIN
MENU screen within 5 seconds without touching any key, it will automaticaly return to analyze function.
If the analyzer is in subfunction mode,
in most cases, it will automaticaly return to analyze mode within 10 minutes.
The Analyze function screen shows the impurity concentration and the
application gases in the first line, and the range in the second line. In the
lower right corner, the abbreviation Anlz indicates that the analyzer is in the
Analyze mode. If there is an * before the Anlz, it indicates that the range is
linearized.
1.95 % H2 N2
nR1:Ø 10 *Anlz
n indicates non inverting range
i indicates inverting range
If the concentration detected is over range, the first line of the display
blinks continuously.
4.8
Programming
CAUTION: The programming functions of the Set Range and
Curve Algorithm screens are configured at the
factory to the users application specification. These
functions should only be reprogrammed by trained,
qualified personnel.
To program, you must:
1. Enter the password, if you are using the analyzer’s password
protection capability.
2. Connect a computer or computer terminal capable of sending an
RS-232 signal to the analyzer RS-232 connector. (See chapter 3
Installation for details). Send the rp command to the analyzer.
Now you will be able to select the APPLICATION and ALGORITHM setup functions.
Teledyne Analytical Instruments
4-23
4 Operation
Model 2020
4.8.1 The Set Application Screen
The Set Application screen allows reprogramming of the three analysis
ranges and the calibration range (including impurity gas, background gas,
low end of range, high end of range, and % or ppm units). Original programming is usually done at the factory according to the customer’s application. It must be done through the RS-232 port using a computer running a
terminal emulation program.
Note: It is important to distinguish between this System programming subfunction and the Range button function, which is an
operator control. The Set Range Screen of the Application
function allows the user to DEFINE the upper and lower limits
of a range AND the application of the range. The Range function only allows the user to select or define the limits, or to
select the application, but not to define the application.
Normally the Model 2020 is factory set to default to manual range
selection, unless it is ordered as a single-application multiple-range unit (in
which case it defaults to autoranging). In either case, autoranging or manual
range selection can be programmed by the user.
In the autoranging mode, the microprocessor automatically responds
to concentration changes by switching ranges for optimum readout sensitivity. If the upper limit of the operating range is reached, the instrument
automatically shifts to the next higher range. If the concentration falls to
below 85% of full scale of the next lower range, the instrument switches to
the lower range. A corresponding shift in the DC concentration output, and
in the range ID outputs, will be noticed.
The autoranging feature can be overridden so that analog output stays
on a fixed range regardless of the contaminant concentration detected. If the
concentration exceeds the upper limit of the range, the DC output will
saturate at 1 V dc (20 mA at the current output).
However, the digital readout and the RS-232 output of the concentration are unaffected by the fixed range. They continue to read beyond the
full-scale setting until amplifier saturation is reached. Below amplifier
saturation, the over range readings are accurate UNLESS the application
uses linearization over the selected range.
To program the ranges, you must first perform the four steps indicated
at the beginning of section 4.8 Programming. You will then be in the MAIN
MENU and selecting application function screen.
Sel rng to set appl:
> Ø1 Ø2 Ø3 CAL <
4-24
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Operation 4
Use the UP/DOWN key to increment/decrement the range number to
01, 02, 03, or CAL, and Enter.
Imp:
FR:Ø
H2 Bck:
TO:1Ø %
N2
Use the UP/DOWN key to increment the respective parameters as
desired.
Use the ENTER to move from Imp: (impurity) to Bck: (background),
FR: (from—lower end of range), TO: (to—upper end of range), and PPM
or %.
Last Enter will accept the values and the screen will display
OFFST/INVRT:
Standard:
ESC if your application is standard. If your application requires
OFFSET or Reversal output, ENTER will set your output as OFFSET or
REVERSAL. (See special function setup section 4.9 for more information).
. (See note below.) Repeat for each range you want to set.
Note: The ranges must be increasing from low to high, for example,
if Range 1 is set to 0–10 % and Range 2 is set to 0–100 %,
then Range 3 cannot be set to 0–50 % since that makes
Range 3 lower than Range 2.
Ranges, alarms, and spans are always set in either percent or ppm
units, as selected by the operator, even though all concentration-data outputs change from ppm to percent when the concentration is above
9999 ppm.
Note: When performing analysis on a fixed range, if the concentration rises above the upper limit as established by the operator
for that particular range, the output saturates at 1 V dc (or 20
mA). However, the digital readout and the RS-232 output
continue to read regardless of the analog output range.
To end the session, send:
st
st
to the analyzer from the computer.
Teledyne Analytical Instruments
4-25
4 Operation
Model 2020
4.8.2 The Curve Algorithm Screen
The Curve Algorithm is a linearization method. It provides from 1 to 9
intermediate points between the ZERO and SPAN values, which can be
normalized during calibration, to ensure a straight-line input/output transfer
function through the analyzer. (Before setting the alogorithm curve, each
range must be Zeroed and Spanned).
Each range is linearized individually, as necessary, since each range
will usually have a totally different linearization requirement.
To linearize the ranges, you must first perform the four steps indicated
at the beginning of section 4.8 Programming. You will then be in the MAIN
MENU and select ALOGORITHM.
4.8.2.1 Checking the linearization
From the MAIN MENU screen, select ALGORITHM, and Enter.
Sel rng set algo
> Ø1 Ø2 Ø3
<
Use the UP/DOWN key to select the range: 01, 02, or 03. Then press
Enter.
Gas use: H2 N2
Range: Ø -10 %
Enter again.
Algorithm setup:
VERIFY
SETUP
UP/DOWN to select and Enter VERIFY to check whether the linearization has been accomplished satisfactorily.
Dpt
Ø
INPUT
Ø.ØØ
OUTPUT
Ø.ØØ
The leftmost digit (under Dpt) is the number of the data point being
monitored. Use the UP/DOWN keys to select the successive points.
The INPUT value is the input to the linearizer. It is the simulated output
of the analyzer. You do not need to actually flow gas.
The OUTPUT value is the output of the linearizer. It should be the
ACTUAL concentration of the span gas being simulated.
If the OUTPUT value shown is not correct, the linearization must be
corrected. ESCAPE to return to the previous screen. Select and Enter SET
UP to Calibration Mode screen.
4-26
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Operation 4
Select algorithm
mode : AUTO
There are two ways to linearize: AUTO and MANUAL: The auto mode
requires as many calibration gases as there will be correction points along
the curve. The user decides on the number of points, based on the precision
required.
The manual mode only requires entering the values for each correction
point into the microprocessor via the front panel buttons. Again, the number
of points required is determined by the user.
Note: Before performing section 4.8.2 or 4.8.2.3, you must check to
ensure that your calibration gases or points are between low
end and high end of the range setup. All correction points
must be between the Zero and the Span concentration. Do
not enter the Zero and Span points as part of the correction
4.8.2.2 Manual Mode Linearization
To linearize manually, you must have previous knowledge of the
nonlinear thermal-conductivity characteristics of your gases. You enter the
value of the differential between the actual concentration and the apparent
concentration (analyzer output). Analytical Instruments has tabular data of
this type for a large number of gases, which it makes available to customers
on request. See Appendix for ordering information. To enter data:
From the MAIN MENU Screen—
1. Use UP/DOWN to select ALGORITHM , and Enter.
2. Select and Enter SETUP.
3. Select MANUAL from the Calibration Mode Select screen.
Dpt
Ø
INPUT
Ø.ØØ
OUTPUT
Ø.ØØ
The data entry screen resembles the verify screen, but the gas values
can be modified and the data-point number cannot.
Use the UP/DOWN key to set the INPUT value for the lowest concentration into the first point. ENTER to move to OUTPUT field. Use the
UP/DOWN key to set the OUTPUT value, the lowest concentration into
the first point. ENTER to accept the first setting.
After each point is entered, the data-point number increments to the
next point. Moving from the lowest to the highest concentration, use the
UP/DOWN keys to set the proper values at each point.
Teledyne Analytical Instruments
4-27
4 Operation
Model 2020
Dpt
0
INPUT
Ø.ØØ
OUTPUT
Ø.ØØ
Repeat the above procedure for each of the data points you are setting
(up to nine points: 0-8). Set the points in unit increments. Do not skip
numbers. The linearizer will automatically adjust for the number of points
entered.
When you are done, ESCAPE. The message, Completed. Wait for
calculation, appears briefly, and then the main System screen returns.
To end the session, send:
st
st
to the analyzer from the computer.
4.8.2.3 Auto Mode Linearization
To linearize in the Auto Mode, you must have on hand a separate
calibration gas for each of the data points you are going use in your linearization. First, the analyzer is zeroed and spanned as usual. Then, each
special calibration gas, for each of the intermediate calibration points, is
flowed, in turn, through the sensor. As each gas flows, the differential value
for that intermediate calibration point is entered from the front panel of the
analyzer.
Note: The span gas use to span the analyzer must be >90% of the
range being analyzed.
Before starting linearization, perform a standard calibration. See
section 4.4. To enter data:
From the MAIN MENU screen—
1. Use UP/DOWN to select ALGORITHM , and Enter.
2. Select and Enter SETUP.
3. Enter AUTO from the Calibration Mode Select screen.
The Auto Linearize Mode data entry screen appears.
1.95 %
H2
Input(Ø) :2.00
N2
5. Use the UP/DOWN keys to set the proper value of calibration
gas, and Enter. Repeat this step for each cal-point number as it
appears in the Input (x) parentheses.
4-28
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Operation 4
6. Repeat step 5 for each of the special calibration gases, from the
lowest to the highest concentrations. Escape when done.
To end the session, send:
st
st
to the analyzer from the computer.
4.9 Special Function Setup
CAUTION: The programming functions of the output signal
reversal, polarity reversal and gain preset are configured at the factory to the users application specification. These functions should only be reprogrammed by trained, qualified personnel.
4.9.1 OFFSET OUTPUT/Reversal Output
4.9.1.1 Output Signal Reversal
Some applications require a reversal of the output signals in order
for the 4-20mA and 0-1 V DC output signals to correspond with the low
and high end of the concentration range. For example, if an application
involves the analysis of 85-100% oxygen in a background of argon by
measuring the thermal conductivity of the binary gas, the analyzer would
normally be set up so that the 100% oxygen (0% argon) concentration
would correspond to the zero level (4mA 0 V) of the output signal. Then,
85% oxygen (15% argon) would correspond to 20mA (1 V) in the signal
output.
It may be convenient for the user to have the outputs reversed so that
the 85-100% oxygen level outputs a 4-20mA (0-1 V) signal respectively.
This can be accomplished by reversing the data input to the custom settings.
Not all applications will require a reversing function, however, if this is
desirable. This can be programmed by allowing the user to set the analyzer
to read reversal output. Contact the factory for further information.
4.9.1.2 Output Signal Offset
TAI has provided the output offset feature in the software for the
accuracy purpose. In many cases, the analyzer does not require this feature.
For exmple, if the analyzer has setup to analyze the sample gas of 40-50%
Teledyne Analytical Instruments
4-29
4 Operation
Model 2020
argon in nitrogen, normally zero gas of this application requires 40% argon
in nitrogen. However, 100% of nitrogen can be used to zero the analyzer.
In this case, the output offset is not needed to setup. For linear output the
accuracy will access successfully within +/-1% off. But, if the application is
analyzing the sample gas that is not linearly, the accuracy of the analyzer
may not meet the specification. Therefore, output offset require and yet
40% argon in nitrogen is also needed for zero calibration gas.
To set up the output reversal or output offset, see section 4.8.1-set
APPLICATION sceen.
NOTE: If the inverting has been setup, “i” shall display on the left
bottom corner. Otherwise, the left bottom corner display ”n”.
4.9.2 Polarity Reversal
In some special applications, user will find that the display dispalys
negative concentration values, even if proper span gas is injected. For
example, if application involves the analysis of 0-10% nitrogen in argon, and
the reference chamber is sealed with air instead of argon, the microprocessor will be detecting signal and process assuming that the reference chamber
is flowing with argon. However, in this case, the seal air reference that is
compared to background of argon from the measurement chamber which
has higher coeffiction parameter and then will cause the analyzer to go
negative. To correct this problem, TAI has added the Polarity Correction
feature. This feature can be set as follow:
Close S1-5 range 1
Close S1-6 range 2
Close S1-7 range 3
Close S1-8 cal range
Select STANDBY to restart the system.
4.9.3 Gain Preset
NOTE: This function will apply only for the analizer that has multiple
range with single application and non-linearity.
4-30
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Operation 4
For nonlinear application, signal that produces from the termal conductivity, will not correspond to the actual gas concentration. The amplification of each range will not agree, therefore, the gain must be preset in order
for the signal to read linearly. To set the gain, the following must be performed in sequence.
NOTE: Before setting up this feature, you must have a span gas
containing 90%-100% of the lowest range of the analyzer.
1. Set unit range to lowest range.
NOTE: For output reversal, the lowest range should be range 3, else
the lowest range is range 1.
2. Connect span gas to span inlet.
3. Use UP/DOWN key to scroll to the CAL-INDEPD function.
4. Hold Escape/Enter control to ENTER position for approximately 10
seconds until the upper right connect display “ok” message.
5. Use UP/DOWN scroll to select SPAN function, then setup the
setting to span level. Press ENTER key to span.
NOTE: You must do step #5 before the analyzer return to analyze
mode. If the analyzer returns to analyze function, you must
repeat step 3-5 again.
6. Set range to nex range’
7. Repeat steps 3-5 until you reach the last range of the row.
Teledyne Analytical Instruments
4-31
4 Operation
4-32
Model 2020
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Maintenance 5
Maintenance
5.1
Routine Maintenance
Aside from normal cleaning and checking for leaks at the gas connections, routine maintenance is limited to replacing fuses, and recalibration.
For recalibration, see Section 4.4 Calibration.
WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS
MANUAL.
5.2
System Self Diagnostic Test
Use UP/DOWN key to enter the SELF-TESY function.
The following failure codes apply:
Table 5-1: Self Test Failure Codes
Power
0
1
2
3
OK
5 V Failure
15 V Failure
Both Failed
Analog
0
1
2
3
OK
DAC A (0–1 V Concentration)
DAC B (0–1 V Range ID)
Both Failed
Preamp
0
1
OK
Zero too high
Teledyne Analytical Instruments
5-1
5 Maintenance
Model 2020
2
3
Amplifier output doesn't match test input
Both Failed
>3
Open gain resistor
Cell
5.3
0
OK
1
Failed (open filament, short to ground, no
power.)
2
Unbalance (deterioration of filaments,
blocked tube)
Fuse Replacement
The 2020 requires two 5 x 20 mm, 4 A, T type (Slow Blow) fuses.
The fuses are located inside the explosion proof housing on the Electrical
Connector Panel, as shown in Figure 5-1. To replace a fuse:
1. Disconnect the Unit from its power source.
2. Place a small screwdriver in the notch in the fuse holder cap,
push in, and rotate 1/4 turn. The cap will pop out a few
millimeters. Pull out the fuse holder cap and fuse, as shown in
Figure 5-1.
Figure 5-1: Removing Fuse Cap and Fuse from Holder
3. Replace fuse by reversing process in step 1.
Remove Power to the instrument before changing fuses.
5-2
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
5.4
Maintenance 5
Major Internal Components
The Cell Compartment and Front Panel PCBs are accessed by unlatching and swinging open the front panel, as described earlier. The
balance of the PCBs are accessed by removing the rear panel retaining
screws and sliding out the entire subassembly. See Figure 5-3, below. The
major electronic components locations are shown in Figure 5-4 (with Cell
Compartment removed for clarity).
WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS
MANUAL.
The 2020 contains the following major components:
• Analysis Section
Cell Compartment
Cell Block
• Power Supply
• Preamp and Motherboard with Microcontroller
• Display Board and Displays
5 digit LED meter
2 line, 20 character, alphanumeric, VFD display
• Rear Panel Board.
5.5 Voltage Selections
There is a switch on the interface board, inside the instrument, that
selects the working voltage between 230/115 VAC.
Voltage Selector Switch
Make sure the switch is in the proper position before
powering the instrument.
230V
115V
Teledyne Analytical Instruments
5-3
5 Maintenance
Model 2020
Figure 5-3: Rear Panel Retaining Screws
Figure 5-4: Locations of Printed Circuit Board Assemblies
5-4
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Maintenance 5
See the drawings in the Drawings section in back of this manual
for details.
5.6
Cell, Heater, and/or Thermistor Replacement
The Thermal Conductivity Cell, with its Heater and Thermistor, is
mounted inside the insulated cell compartment, just behind the analyzer's
front panel access door. To remove the one of these components, you must
first slide the entire Cell Compartment out of the analyzer through the front
panel access door as described in the procedure below. Figure 5-5 identifies the five screws that must be removed in order to remove the Cell
Compartment.
5.6.1 Removing the Cell Compartment
WARNING: IF THE MODEL 2020 ANALYZER HAS BEEN USED
WITH TOXIC GASES, FLUSH IT THOROUGHLY
BEFORE PERFORMING THIS PROCEDURE.
WARNING: DISCONNECT ALL POWER TO THE MODEL 2020
BEFORE PERFORMING THIS PROCEDURE.
Figure 5-5: Location of Cell Compartment
Teledyne Analytical Instruments
5-5
5 Maintenance
Model 2020
To remove the Cell Compartment:
a. Disconnect gas and electrical connections to the analyzer.
b. Remove analyzer from its mounting, and remove gas fittings
from the gas ports on the bottom of the analyzer, so that nothing
projects from the ports.
c. Remove the Cell Compartment retaining screws identified in
Figure 5-5. You will have to unlatch and swing open the front
panel door to remove the front screws.
d. Carefully pull the Cell Compartment out through the front of
the analyzer. There is enough length to the cell's electrical
wiring to allow this.
e. After replacing the necessary component and reassembling the
Cell Compartment, Replace the Compartment by reversing the
above procedure, steps a through d.
5.6.2 Removing and Replacing the Cell Block
a. Refer to Figure 5-6, which illustrates removal of the Cell Block
from the Cell Compartment. Exploded view is as seen from the
top of the Cell Block.
Figure 5-6: Removal of Cell from Cell Housing
5-6
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Maintenance 5
b. Remove the two screws holding the front mounting bracket—
they also hold the Cell Block Cover to the Cell Block—and
then pull off the cover.
c. Turn the uncovered Cell Block assembly over so that the
bottom faces you. The black rectangular block with four screws
is the Heater Block. Separate the Heater Block from the Cell
Block by removing the four screws. Leave the Heater Block
electrical connections connected.
d. Remove the four screws from each of the black plates that hold
the Cell. The Cell is sandwiched between the plates. You should
now be able to slide the Cell free.
e. Leave the electrical connections connected at the Cell. Unlace
the cabling, and unplug the grey Cell cable at the Preamplifier
PCB connector, J3. (See Figure 5-4, and/or drawings at the rear
of this manual.) The Preamplifier PCB can be more easily
accessed by removing the analyzer's rear panel. (See Section
5.5.)
f. Replace the cell by reversing the above procedure, steps a
through e.
5.6.3 Removing the Heater and/or Thermocouple
a. Refer to Figure 5-7, which illustrates removal of the Thermistor
and/or Heater from the Cell Compartment. Exploded view is as
seen from the bottom of the Cell Block.
Figure 5-7: Removing the Heater and/or Thermocouple
b. Remove the two screws holding the front mounting bracket—
they also hold the Cell Block Cover to the Cell Block—and
then pull off the cover.
Teledyne Analytical Instruments
5-7
5 Maintenance
Model 2020
c. Turn the uncovered Cell Block assembly over so that the
bottom faces you. The black rectangular block with four screws
is the Heater Block.
d. The Heater is fastened to the Heater Block by a set screw as
well as the silicone sealing compound. The Thermistor is
fastened only by the silicone sealer.
(1) To remove the Heater, use a 1/16 ″ Allen wrench to loosen
the Thermistor set screw. Then, grasp BOTH Heater wires
firmly, and pull the Heater slowly out of the Heater Block,
breaking the silicone seal. Do not allow any foreign matter
to enter the empty duct.
(2) To remove the Thermistor, grasp BOTH Thermistor wires
firmly, and pull the Thermistor slowly out of the Heater
Block, breaking the silicone seal. Do not allow any foreign
matter to enter the empty duct.
e. Undo the cable lacing and separate the Heater/Thermistor wires.
Then, disconnect the wires from TS1 on the Temperature
Control Board. (See Figures 5-4 and 5-7.)
5.6.4 Replacing the Heater and/or Thermocouple
a. To replace the Heater and/or Thermocouple, coat the new
element with silicone sealing compound, and insert it into the
duct.
CAUTION: The larger duct is for the Heater element, and the
smaller duct is for the Thermocouple.
b. Enough sealing compound should be on the element to spill
over and seal around the wire where it enters the duct. Smooth
the outer seal and remove any excess.
c. Reassemble the Cell Compartment by reversing the procedure
in section 5.6.3. Then replace the cabling.
d. Reinstall the Assembled Cell Compartment by reversing the
procedure in section 5.6.1. Then reconnect the wires to TS1 on
the Temperature Control board.
5-8
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
5.7
Maintenance 5
Cleaning
If instrument is unmounted at time of cleaning, disconnect the instrument from the power source. Close and latch the front-panel access door.
Clean outside surfaces with a soft cloth dampened slightly with plain clean
water. Do not use any harsh solvents such as paint thinner or benzene.
For mounted instruments, DO NOT wipe the front panel while the
instrument is controlling your process. Clean the front panel as prescribed
in the above paragraph.
5.8
Phone Numbers
Customer Service: (818) 934-1673
Environmental Health and Safety: (818) 934-1592
Fax: (818) 961-2538
EMERGENCY ONLY: (24-hour pager) 1-800-759-7243
PIN # 1858192
Teledyne Analytical Instruments
5-9
5 Maintenance
5-10
Model 2020
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Appendix
Appendix
A-1 Specifications
Ranges: Three ranges plus a cal range with linearizer,
field selectable within specified limits (application dependent) and Auto Ranging
Display: 2 line by 20 alphanumeric VFD accompanied
by 5 digit LED display
Accuracy: ±1% of full scale for most binary mixtures at
constant temperature
±5% of full scale over operating temperature
range once temperature equilibrium has been
reached
Response Time: 90% in less than 65 seconds
System Operating
Temperature: 32°F to 122°F (0 - 50°C)
Sensor Type: Standard TC cell filaments detector)
Signal Output: Two 0-1 VDC (concentration and range ID)
Two 4-20 mADC isolated (concentration and
range ID)
Alarm: Two fully programmable concentration alarm
set points and corresponding Form C, 3 amp
contacts.
One system failure alarm contact to detect
power, calibration, zero / span and sensor
failure.
Teledyne Analytical Instruments
A-1
Appendix
Models 2020
System Power
Requirements: 115/230 VAC, 50-60Hz
Cell Material: Nickel plated brass block with nickel alloy
filaments and stainless steel and plates
O/P Interface: Full duplex RS-232, implement a subset of
Tracs Command
Mounting: Explosion-Proof Housing, Bulkhead Mounting.
Options: (C) Electrically operated CAL/ZERO valves
(H) SS Sampling System (SS cell block gold
filaments)
(R) Sealed Reference TC Cell
(L) Flow Control Gas Panel
(F) Flame Arrestors for group C & D service
with Flow Control Gas Panel.
(O) Flame Arrestors for group B service with
Control Gas Panel.
(P) Flame Arrestors for group C & D with
CAL valves with Control Gas Panel.
(Q) Flame arrestors for group B with CAL
valves with Control Gas Panel.
A-2
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Appendix
A-2 Recommended 2-Year Spare Parts List
Qty
Part Number Description
1
D67472
Back Panel Board
1
C62371B
Front Panel Board
1
C65098A
Preamplifier Board
1
C62365D
Main Computer Board
2
F1295
Fuse, 4 A, 250 V, 5 × 20 mm, T—Slow Blow
_____________________
* Order one type only: US or European, as appropriate.
Note:
Orders for replacement parts should include the part number (if
available) and the model and serial number of the instrument for
which the parts are intended.
Orders should be sent to:
TELEDYNE Analytical Instruments
16830 Chestnut Street
City of Industry, CA 91749-1580
Phone 626) 934-1500, Fax (818) 961-2538
TWX (910) 584-1887 TDYANYL COID
Web:
www.teledyne-ai.com
or your local representative.
A-3 Drawing List
D67110
Outline Drawing (Basic and Standard Options)
D-67113
Interconnection and Piping Diagram
Teledyne Analytical Instruments
A-3
Appendix
A-4
Models 2020
Teledyne Analytical Instruments