Endress + Hauser Operating Instructions Oxymax COS61 Dissolved oxygen sensor
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Operating Instructions
Oxymax COS61
Dissolved oxygen sensor
BA00387C/07/EN/13.12
71162645
Documentation information
Warnings
The structure, signal words and safety colors of the signs comply with the specifications of
ANSI Z535.6 ("Product safety information in product manuals, instructions and other collateral
materials").
Safety message structure
Meaning
This symbol alerts you to a dangerous situation.
! DANGER
Failure to avoid the situation will result in a fatal or serious
Cause (/consequences)
Consequences if safety message injury.
is not heeded
► Corrective action
This symbol alerts you to a dangerous situation.
! WARNING
Failure to avoid the situation can result in a fatal or serious
Cause (/consequences)
Consequences if safety message injury.
is not heeded
► Corrective action
This symbol alerts you to a dangerous situation.
! CAUTION
Failure to avoid this situation can result in minor or medium
Cause (/consequences)
Consequences if safety message injury.
is not heeded
► Corrective action
This symbol alerts you to situations that can result in damage to
NOTICE
property and equipment.
Cause/situation
Consequences if safety message
is not heeded
► Action/note
Symbols used
Èä1
This symbol indicates a cross reference to a defined page (e.g. p. 1).
È å 2 This symbol indicates a cross reference to a defined figure (e.g. fig. 2).
Additional information, tips
Permitted or recommended
Forbidden or not recommended
Endress+Hauser
Oxymax COS61
Table of contents
1
Basic safety instructions . . . . . . . . . . . . 4
10
Technical data . . . . . . . . . . . . . . . . . . . 28
1.1
1.2
1.3
1.4
1.5
Requirements for personnel . . . . . . . . . . . . . . . . . . .
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Occupational safety . . . . . . . . . . . . . . . . . . . . . . . . .
Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . .
Product safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10.1
10.2
10.3
10.4
10.5
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance characteristics . . . . . . . . . . . . . . . . . .
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Mechanical construction . . . . . . . . . . . . . . . . . . . .
2
Identification . . . . . . . . . . . . . . . . . . . . 6
2.1
2.2
2.3
Product page and configurator . . . . . . . . . . . . . . . . . 6
Order code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3
Installation . . . . . . . . . . . . . . . . . . . . . . 7
3.1
3.2
3.3
3.4
3.5
Incoming acceptance, transport, storage . . . . . . . . . . 7
Installation conditions . . . . . . . . . . . . . . . . . . . . . . . 7
Installation instructions . . . . . . . . . . . . . . . . . . . . . . 8
Installation examples . . . . . . . . . . . . . . . . . . . . . . . 10
Post-installation check . . . . . . . . . . . . . . . . . . . . . . 14
4
Wiring . . . . . . . . . . . . . . . . . . . . . . . . 15
4.1
4.2
4.3
Direct connection to the transmitter . . . . . . . . . . . 15
Connection via junction box . . . . . . . . . . . . . . . . . 16
Post-connection check . . . . . . . . . . . . . . . . . . . . . . 16
5
Device description . . . . . . . . . . . . . . . 17
5.1
5.2
5.3
Sensor design . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . 18
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
6
Commissioning. . . . . . . . . . . . . . . . . . 21
6.1
6.2
6.3
Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Automatic cleaning . . . . . . . . . . . . . . . . . . . . . . . . 21
7
Maintenance. . . . . . . . . . . . . . . . . . . . 22
7.1
7.2
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Replacing wear and tear materials . . . . . . . . . . . . . 23
8
Accessories. . . . . . . . . . . . . . . . . . . . . 24
8.1
8.2
8.3
Connection accessories . . . . . . . . . . . . . . . . . . . . . 24
Installation accessories . . . . . . . . . . . . . . . . . . . . . . 24
Cleaning and calibration . . . . . . . . . . . . . . . . . . . . 25
9
Trouble-shooting . . . . . . . . . . . . . . . . 26
9.1
9.2
9.3
9.4
9.5
Trouble-shooting instructions . . . . . . . . . . . . . . . .
Sensor checks . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Endress+Hauser
4
4
4
4
5
28
28
28
28
29
Index . . . . . . . . . . . . . . . . . . . . . . . . . . 30
26
26
27
27
27
3
Basic safety instructions
Oxymax COS61
1
Basic safety instructions
1.1
Requirements for personnel
► Installation, commissioning, operation and maintenance of the measuring system must only be
carried out by trained technical personnel.
► The technical personnel must be authorized by the plant operator to carry out the specified
activities.
► The electrical connection may only be performed by an electrical technician.
► The technical personnel must have read and understood these Operating Instructions and must
follow the instructions they contain.
► Measuring point faults may only be rectified by authorized and specially trained personnel.
Repairs not described in the enclosed Operating Instructions may only be carried out directly
at the manufacturer's or by the service organization.
1.2
Designated use
The oxygen sensor is suitable for continuous measurement of dissolved oxygen in water.
Typical applications are:
• Measuring, monitoring and regulating the oxygen content in activated sludge basins.
• Monitoring the oxygen content in the sewage treatment plant outlet.
• Monitoring, measuring and regulating the oxygen content in public waters and fish farming
water.
• Monitoring of oxygen enrichment in drinking water.
Any other use than the one described here compromises the safety of persons and the entire
measuring system and is not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.
1.3
Occupational safety
As the user, you are responsible for complying with the following safety conditions:
• Installation instructions
• Local standards and regulations
Electromagnetic compatibility
With regard to electromagnetic compatibility, this device has been tested in accordance with the
applicable European standards for industrial applications.
The electromagnetic compatibility indicated only applies to a device that has been connected in
accordance with the instructions in these Operating Instructions.
1.4
Operational safety
► Before commissioning the entire measuring point, make sure all the connections are correct.
Ensure that electrical cables and hose connections are not damaged.
► Do not operate damaged products, and safeguard them to ensure that they are not operated
inadvertently. Mark the damaged product as defective.
► If faults cannot be rectified, the products must be taken out of service and secured against
unintentional commissioning.
CAUTION
The cleaning system is not switched off during calibration or maintenance activities
Risk of injury due to medium or cleaning agent
► If a cleaning system is connected, switch it off before removing a sensor from the medium.
► If you are not switching off the cleaning system because you wish to test the cleaning function,
wear protective clothing, goggles and gloves or take other appropriate measures.
!
4
Endress+Hauser
Oxymax COS61
Basic safety instructions
1.5
Product safety
The product is designed to meet state-of-the-art safety requirements, has been tested and left the
factory in a condition in which it is safe to operate. Relevant regulations and European standards
have been observed.
Endress+Hauser
5
Identification
Oxymax COS61
2
Identification
2.1
Product page and configurator
You can create a complete and valid order code by using the configurator on the internet product
page.
Product page link:
www.products.endress.com/cos61
2.2
Order code
1.
You can choose from the following options on the product page located on the right:
2.
Click "Configure this product".
3.
The configurator opens in a separate window.
Use the radio buttons to configure the order code from the nameplate of your device.
4.
Afterwards, you can export the order code as a PDF or Excel file.
To do so, click the appropriate button at the top of the page.
2.3
Scope of delivery
The following items are included in the delivery:
• Oxygen sensor with transport protection cap for membrane protection
• Operating Instructions (on CD only)
• Brief Operating Instructions (paper version)
If you have any questions, please contact your supplier or your local sales center.
6
Endress+Hauser
Oxymax COS61
Installation
3
Installation
3.1
Incoming acceptance, transport, storage
► Make sure the packaging is undamaged!
► Inform the supplier about any damage to the packaging.
Keep the damaged packaging until the matter has been settled.
► Make sure the contents are undamaged!
► Inform the supplier about damage to the contents. Keep the damaged products until the matter
has been settled.
► Check that the order is complete and agrees with your shipping documents.
► The packaging material used to store or to transport the product must provide shock protection
and humidity protection. The original packaging offers the best protection. Also, keep to the
approved ambient conditions (see "Technical data").
► If you have any questions, please contact your supplier or your local sales center.
3.2
Installation conditions
3.2.1
Dimensions
a0004083-en
Fig. 1:
Endress+Hauser
Fixed cable version
a0004084-en
Fig. 2:
TOP68 version
7
Installation
Oxymax COS61
3.2.2
Angle of installation
The sensor can be installed up to the horizontal in an assembly, support or a suitable process
connection.
Other angles and overhead installation are not recommended. Reason: possible sediment formation
and resulting falsification of measured value.
The optimum installation angle is 45˚ (e.g. with assembly CYA112).
a0004087
Fig. 3:
A
Angle of installation
Permissible installation positions: 0 ... 180 ˚, optimum angle 45 ˚
► Make sure you comply with the instructions for installing sensors. You will find them in the
Operating Instructions for the assembly used.
3.2.3
Mounting location
• Select the installation location so that there is easy access for later calibration.
• Make sure that upright posts and assemblies are secured safely and vibration-free.
• Select an installation location which produces a typical oxygen concentration.
3.3
Installation instructions
3.3.1
Measuring system
A complete measuring system comprises:
• Oxygen sensor Oxymax COS61
• Transmitter, e.g. Liquisys COM2x3-W
• Special measuring cable
• Assembly, e.g. COA250 flow assembly, CYA112 immersion assembly or COA451 retractable
assembly
Optional:
• CYH112 assembly holder for immersion operation
• VS junction box (for cable extension)
• Cleaning system
8
Endress+Hauser
Oxymax COS61
Installation
a0012885
Fig. 4:
1
2
3
Measuring system (example)
Liquisys COM253 transmitter
COS61 oxygen sensor
Pipework (ascending pipe)
3.3.2
4
5
6
Medium flow direction
Retractable assembly Cleanfit COA451
Sensor cable
Installing a measuring point
For immersed operation, install the individual modules away from the basin on a solid base.
Only carry out the final installation at the intended installation location.
For a complete installation of a measuring point, proceed as follows:
1.
Install a retractable or a flow assembly (if used) into the process.
2.
Connect the water supply to the rinse connections (if you use an assembly with cleaning
function).
3.
Install and connect the oxygen sensor.
4.
Install an immersion or an suspension assembly (if used) into the process.
NOTICE
No assembly used, sensor not correctly installed, grounding regulations not observed
Risk of damaging the sensor cable, no protection to electromagnetic interferencies
► Screw the sensor into the assembly so that the cable is not twisted.
► Avoid exerting excessive tensile force on the cable (e.g. from jerky pulling).
► When using metallic assemblies and installation equipment, comply with national grounding
regulations.
► Observe the sensor installation instructions of the Operating Instructions of the assembly used.
Endress+Hauser
9
Installation
Oxymax COS61
3.4
Installation examples
3.4.1
Immersion operation
Upright post and chain assembly
For large basins, where sufficient installation distance is required from the basin edge (aeration
basin, especially), it is advisable to use the upright post and chain assembly. The free swinging of
the immersed assembly practically rules out vibrations from the upright post.
A good self-cleaning of the fluorescence cap is reached due to the swinging of the assembly.
According to this effect, the sensor life time can be extended.
a0013510
Fig. 5:
1
2
3
4
5
10
Chain holder, rail mounted
Chain
Flexdip CYH112 holder
Rail
Oxymax sensor
Flexdip CYA112 wastewater assembly
a0013511
Fig. 6:
1
2
3
4
5
6
Chain holder, mounted to a post
Weather protection cover
#ontroller / transmitter
Chain
Flexdip CYA112 wastewater assembly
Oxymax sensor
Flexdip CYH112 holder
Endress+Hauser
Oxymax COS61
Installation
Upright post and fixed immersion assembly
The preferable type of installation for strong or turbulent flow (> 0.5 m/s) in the basin or open
channels is to secure the device to an upright post and a securely mounted immersion tube. If the
flow is very strong, a second transverse pipe can be installed with its own pipe support.
a0013513
Fig. 7:
1
2
3
4
5
Endress+Hauser
Assembly holder with immersion tube
Weather protection cover CYY101
Controller / transmitter
Immersion assembly Flexdip CYA112
Oxymax sensor
Assembly holder Flexdip CYH112
11
Installation
Oxymax COS61
Basin rim mounting with immersion assembly
For fixing to the sides of the basin or channel, we recommend the pendulum holder of the
immersion tube. Optionally, you can also use the assembly with a float.
1
2
3
4
a0013693
Fig. 8:
1
2
3
4
Basin rim mounting
Pendulum holder CYH112
Assembly Flexdip CYA112
Float of assembly CYA112
Oxymax sensor
Floating body
To aid installation in strongly fluctuating water levels, e.g. in rivers or lakes, there is a floating body
COA 110-50 available (→ å 9).
mm/inch
1
2
1
500/19.69
800/31.50
220/8.66
2
Ø200/7.87
3
4
5
6
7
3
Cable route with strain relief and rain protection
Mounting ring for ropes and chains with locking
screw
Lugs Ø15, 3 x 120 ˚ for anchoring
Saltwater-resistant plastic float
Pipe 40x1, stainless steel 1.4571 (AISI 316Ti)
Shock absorber and weight
Oxygen sensor
4
G1
5
6
7
a0004108-en
Fig. 9:
12
Floating body
Endress+Hauser
Oxymax COS61
Installation
3.4.2
Flow assembly
The COA250 flow assembly with automatic self-venting is suitable for use in pipelines or hose
connections. The inlet is at the bottom of the assembly, the outlet at the top (connection thread
G¾). It can be installed in a pipe by using two 90˚ pipe brackets to allow inflow to the assembly
(→ å 11, Pos. 6).
1
8
3
7
6
5
2
4
a0004114
Fig. 11:
a0004113-en
Fig. 10:
1
2
3
4
5
Flow assembly COA250
Screw-in part for sensor
Screw ring
Meter body
Connection thread G¾
Dummy plug (connection for spray head CUR3)
3.4.3
1
2
3
4, 7
5
6
8
Bypass installation with manually
actuated valves or solenoid valves
Main line
Medium return
Oxygen sensor
Manually actuated or solenoid valves
Flow assembly COA250
90 ˚ pipe bracket
Medium removal
Retractable assembly
The assembly is designed for installation on tanks and pipes. Suitable nozzles must be available for
this.
Install the assembly at places with constant flow. The minimum pipe diameter is DN 80 (3").
a0004117-en
Fig. 12:
1
2
3
4
5
Endress+Hauser
Permissible and impermissible sensor installation positions
Ascending pipe, best position
Horizontal pipe, sensor top down, impermissible due to air cushion or foam bubble forming
Horizontal pipe, installation with permissible emitting angles (acc. to sensor version)
Overhead installation, impermissible due to missing electrolyte contact of the sensor electrodes
Down pipe, impermissible
13
Installation
Oxymax COS61
NOTICE
Sensor not immersed into the medium, suspended particles settled on the sensor
membrane or sensor optics, sensor installed overhead
Measuring errors can occur
► Do not install the assembly at places, where air cushions or foam bubbles can be formed or
where suspended particles can settle on the sensor membrane or optics (→ å 12).
3.5
►
►
►
►
►
14
Post-installation check
Sensor and cable undamaged?
Cap undamaged?
Compliance with permissible sensor installation position?
Is the sensor installed in an assembly and is not suspended from the cable?
Avoid moisture by rain by putting the protective cap on the assembly?
Endress+Hauser
Oxymax COS61
Wiring
4
Wiring
WARNING
Device is energized
Improper connection can cause injury or death.
► The electrical connection must only be carried out by a certified electrician.
► Technical personnel must have read and understood the instructions in this manual and must
adhere to them.
► Prior to beginning any wiring work, make sure voltage is not applied to any of the cables.
!
4.1
Direct connection to the transmitter
4.1.1
Field installation
Connect the sensor directly to the transmitter by using the special measuring cable with SXP plug.
a0004118
Fig. 13:
4.1.2
SXP plug
Panel mounting
• Remove the SXP connector (transmitter side!) from the cable.
• Refer to the following table for the cable assignment and the assigned terminals for
Liquisys COM223-WX/WS.
• Please note that the cable assignment varies depending on the sensor version (fixed cable or
TOP68 connection).
Terminal
COM223
Endress+Hauser
Sensor with fixed cable (OMK)
Sensor with TOP68 connection
(CYK71)
Core
Assignment
Core
Assignment
87
YE
+UB
YE
+UB
0
GY
0V
WH
0V
96
PK
Com. (digital)
GN
Communication (digital)
97
BU
Com. (digital)
BN
Communication (digital)
88
BN
–UB
Koax innen –UB
15
Wiring
Oxymax COS61
4.2
Connection via junction box
To lengthen the sensor connection beyond the length of the fixed cable, you require a junction box
VS.
Always connect the sensor cable with the SXP plug to the junction box.
The cable extension to the transmitter then depends from the transmitter version, i.e. field device
or panel mounted device.
1
82 / 3.23
80 / 3.15
4
YE
GY
PK
BU
BN
2
GN
3
2
WH
mm / inch
1
2
4.3
0
96
97
88
19
18
a0004130
a0004129-en
Fig. 14:
87
Junction box VS to a field device
SXP plug to field device
SXP plug from sensor
Fig. 15:
2
3
4
Junction box VS to a panel mounted device
SXP plug from sensor
Measuring cable (OMK) to the transmitter
Connection department of the transmitter
Post-connection check
Instrument status and specifications
Remarks
Are the sensor, assembly, junction box or cable damaged?
Visual inspection
Electrical connection
Remarks
Does the supply voltage of the transmitter match the specifications on the
nameplate?
Are the installed cables strain-relieved and not twisted ?
Is the cable type route completely isolated ?
Power cable/weak current cable
Are the power supply and signal cable correctly connected to the
transmitter ?
Use the connection diagram of the
transmitter.
Long enough length of cable core stripped and correct in terminal?
Check seating (pull slightly)
Are all the screws terminals properly tightened ?
Tighten
Are all the cable entries installed, tightened and sealed ?
For cable entries lateral: cable loops
downwards for water to be able to drip off.
Are all the cable entries installed downwards or lateral ?
16
Endress+Hauser
Oxymax COS61
Device description
5
Device description
5.1
Sensor design
1
5
2
6
3
7
4
8
a0004151
Fig. 16:
1
2
3
4
5
6
7
8
Sensor design
Sensor cable
Sensor shaft
O-ring
Protection guard
Threaded connection
Detector
Emitter diode
Fluorescence cap
The sensor consists of the following function units:
• Sensor shaft
• Sensor head with optics (emitter and detector)
• Fluorescence cap
• Protection basket
• Alternatively to the protection basket, you can use a spray head COR 3 (optional, see
"Accessories") for use in immersed operation with cleaning function.
Endress+Hauser
17
Device description
Oxymax COS61
5.2
Measuring principle
5.2.1
Oxygen measurement based on the principle of fluorescence
quenching
• Sensor design:
– Oxygen-sensitive molecules (markers) are integrated in an optically active layer (fluorescence
layer).
– The surface of the fluorescence layer is in contact with the medium.
– The sensor optics are directed at the underside of the fluorescence layer.
• There is an equilibrium between the oxygen partial pressure in the medium and that in the
fluorescence layer:
– If the sensor is immersed in the medium, the equilibrium is established very quickly.
• Measuring process:
– The sensor optics send green light pulses to the fluorescence layer.
– The markers "answer" (fluoresce) with red light pulses.
– The duration and intensity of the response signals is directly dependent on the oxygen contents
and the partial pressure.
– If the medium is free from oxygen, the response signals are long and very intense.
– Oxygen molecules quench the marker molecules. As a result, the response signals are shorter
and less intense.
• Measurement result:
– The sensor returns a signal that is in proportion to the oxygen concentration in the medium.
– The medium temperature and air pressure are already taken into account calculated in the
sensor.
– In addition to the standard values of concentration, saturation index and partial pressure, the
sensor also returns a raw measured value in μs. The value corresponds to the decay time of the
fluorescence and is approx. 20 μs in air, and approx. 60 μs in media free from oxygen.
5.2.2
Fluorescence cap
The oxygen dissolved in the medium is diffused into the fluorescence cap.
Suitable flow is not necessarily mandatory but it does improve the speed at which the measuring
system responds and ensures a more representative measured value compared to a measurement in
static medium.
The cap is only permeable for dissolved gases. Other substances dissolved in the liquid phase e.g.
ionic substances, will not penetrate through the membrane. Therefore, medium conductivity has
no impact on the measuring signal.
5.3
Calibration
Calibration is a means of adapting the transmitter to the characteristic values of the sensor.
Normally, sensor calibration is seldom necessary. It is necessary after:
• Changing the fluorescence cap
Within the framework of system monitoring and supervision, for example, the calibration can also
be cyclically monitored (at typical time intervals, depending on operating experience) or renewed.
Ideally, use the calibration vessel (see Accessories) for calibration. For this purpose, unscrew the
basket protector from the sensor and guide the sensor into the calibration vessel as far as it will
go (resting on the vessel rim).
18
Endress+Hauser
Oxymax COS61
Device description
5.3.1
Types of calibration
Types of calibration:
• Air (preferably saturated water vapor, e.g. near the water surface)
– Measured values between 70 and 130 %SAT result in the calibration of the measured value at
air
– Measured values smaller than 15 %SAT result in the calibration of the zero point
• Air-saturated water
– Like air calibration
• Reference measured value (entry at transmitter, sensor remains in the medium).
– Measured values between 50 and 150 %SAT result in the calibration of the measured value to
the reference value while maintaining the zero point
– Measured values smaller than 10 %SAT result in the calibration of the measured value to the
reference value while maintaining the measured value at air
If necessary, calibrate COS61 practically:
• In the air (water-vapor saturated) type of calibration in order to calibrate the measured value at
air.
• In the air-saturated water type of calibration, but while using oxygen-free water, (see "Sensor
check" section), to calibrate the zero point.
5.3.2
Calibration intervals
You can determine the intervals with the following method:
1.
Check the sensor one month after its being put into operation:
– Remove the sensor from the medium.
– Clean the outside of the sensor with a damp cloth.
– Measure the oxygen saturation index at air after 20 minutes.
2.
Decide using the results:
a. If the measured value is not at 100 ±2 %SAT, you have to calibrate the sensor.
b. Otherwise, lengthen the time to the next inspection.
3.
Proceed as per Point 1 after two, four and/or eight months. In this way, you can determine
the optimum calibration interval for your sensor.
5.3.3
Calculation example for the calibration value
As a check, you can calculate the expected calibration value (transmitter display) as shown in the
following example (salinity is 0).
Endress+Hauser
1.
Determine:
– The ambient temperature for the sensor (air temperature for "air" calibration method,water
temperature for "air-saturated water" calibration type)
– the altitude above sea level
– the current air pressure L (=rel. air pressure to sea level) at the time of calibration. (If
undeterminable, use 1013 hPa (407 inH2O) for an approximate calculation.)
2.
Define:
– the saturation value S acc. to the first table
– the factor K acc. to the second table
19
Device description
Oxymax COS61
˚ C / ˚F
S [mg/
l=ppm]
˚ C / ˚F
S [mg/
l=ppm]
˚ C / ˚F
S [mg/
l=ppm]
˚ C / ˚F
S [mg/
l=ppm]
0 / 32
14.64
11 / 52
10.99
21 / 70
8.90
31 / 88
7.42
1 / 34
14.23
12 / 54
10.75
22 / 72
8.73
32 / 90
7.30
2 / 36
13.83
13 / 55
10.51
23 / 73
8.57
33 / 91
7.18
3 / 37
13.45
14 / 57
10.28
24 / 75
8.41
34 / 93
7.06
4 / 39
13.09
15 / 59
10.06
25 / 77
8.25
35 / 95
6.94
5 / 41
12.75
16 / 61
9.85
26 / 79
8.11
36 / 97
6.83
6 / 43
12.42
17 / 63
9.64
27 / 81
7.96
37 / 99
6.72
7 / 45
12.11
18 / 64
9.45
28 / 82
7.82
38 / 100
6.61
8 / 46
11.81
19 / 66
9.26
29 / 84
7.69
39 / 102
6.51
9 / 48
11.53
20 / 68
9.08
30 / 86
7.55
40 / 104
6.41
10 / 50
11.25
Altitude
[m / ft]
K
Altitude
[m / ft]
K
Altitude
[m / ft]
K
Altitude
[m / ft]
K
0
1.000
550 / 1800
0.938
1050 / 3450
0.885
1550 / 5090
0.834
50 / 160
0.994
600 / 1980
0.932
1100 / 3610
0.879
1600 / 5250
0.830
100 / 330
0.988
650 / 2130
0.927
1150 / 3770
0.874
1650 / 5410
0.825
150 / 490
0.982
700 / 2300
0.922
1200 / 3940
0.869
1700 / 5580
0.820
200 / 660
0.977
750 / 2460
0.916
1250 / 4100
0.864
1750 / 5740
0.815
250 / 820
0.971
800 / 2620
0.911
1300 / 4270
0.859
1800 / 5910
0.810
300 / 980
0.966
850 / 2790
0.905
1350 / 4430
0.854
1850 / 6070
0.805
350 / 1150
0.960
900 / 2950
0.900
1400 / 4600
0.849
1900 / 6230
0.801
400 / 1320
0.954
950 / 3120
0.895
1450 / 4760
0.844
1950 / 6400
0.796
450 / 1480
0.949
1000 / 3300
0.890
1500 / 4920
0.839
2000 / 6560
0.792
500 / 1650
0.943
3.
Calculate the factor L:
relative air pressure during calibration
L=
-----------------------------------------------------------------------------
1013 hPa
4.
Calculate the calibration value C:
C=S.K.L
Example
• Air calibration at 18˚C (64 ˚F), altitude 500 m (1650 ft) above sea level, air pressure 1009 hPa
(405 inH2O)
• S = 9.45 mg/l, K = 0.943, L = 0.996
Calibration value C = 8.88 mg/l.
You do not need factor K from the table if your device returns the absolute air pressure Labs
(location-dependent air pressure) as the measured value.
Thus, the formula for calculation is: C = S . Labs.
20
Endress+Hauser
Oxymax COS61
Commissioning
6
Commissioning
6.1
Function check
Before first commissioning, check if:
• the sensor is correctly installed
• the electrical connection is correct.
If using an assembly with automatic cleaning, check the correct connection of the cleaning agent
(e.g. water or air).
! WARNING
Escaping process medium
Risk of injury from high pressure, high temperatures or chemical hazards
► Before applying compressed air to an assembly with cleaning facility, make sure the connections
are correctly fitted.
► Do not install the assembly in the process if you cannot make the correct connection reliably.
6.2
Calibration
The sensor is calibrated at the factory. A new calibration is only needed in special situations.
6.3
Automatic cleaning
Compressed air is most suitable for cyclic cleaning. The cleaning unit is either ready supplied or can
be retrofitted, and is attached to the sensor head. It operates at a capacity of 20-60 l/min. Optimum
results are achieved with 2 bar (29 psi) and 60 l/min.
The following settings are recommended for the cleaning unit:
Type of soiling
Endress+Hauser
Cleaning interval
Cleaning duration
Media containing grease and oils
15 min
20 s
Biofilm
60 min
20 s
21
Maintenance
Oxymax COS61
7
Maintenance
Maintenance work must carried out at regular intervals. To ensure that it is carried out, we
recommend you enter the maintenance dates into an operations logbook or in an operations
calendar in advance.
The maintenance cycle primarily depends on:
• the system
• the installation conditions and
• the medium in which measurement is taking place.
The following activities must be carried out:
• Cleaning the sensor
• If necessary, replacing wear and tear materials:
– sealing ring
– electrolyte
– fluorescence cap
• Check the measuring function:
– Remove the sensor from the medium.
– Clean and dry the membrane.
– After about 10 minutes, measure the oxygen saturation index in air (without recalibration).
– The measured value should be near to 100% SAT
(display of O2 saturation with COM 2x3: press plus key four times).
• Recalibration.
(if desired or required)
7.1
Cleaning
The measurement can be corrupted by sensor fouling or malfunction, e.g.:
• Buildup on the fluorescence cap
--> cause longer response times and a reduced slope under certain circumstances.
To ensure reliable measurement, the sensor must be cleaned at regular intervals. The frequency and
intensity of the cleaning operation depend on the measuring medium.
7.1.1
External cleaning
Clean the outside of the sensor:
• before every calibration
• at regular intervals during operation as necessary
• before returning it for repairs.
Depending on the type of soiling, proceed as follows:
Type of soiling
Cleaning
Salt deposits
Immerse the sensor in drinking water or in 1-5% hydrochloric acid
for a few minutes. Afterwards, rinse it with copious amounts of water.
Dirt particles on the sensor body
(not cap!)
Clean the sensor body mechanically with water and a suitable brush.
Dirt particles on the fluorescence cap
Clean with water and a soft sponge.
► After cleaning, rinse the sensor with copious amounts of clean water.
22
Endress+Hauser
Oxymax COS61
Maintenance
7.1.2
Cleaning the optics
The optics only need to be cleaned if medium has penetrated through a defective fluorescence cap.
To clean it, proceed as follows:
1.
Unscrew the protection guard and fluorescence cap from the sensor head.
2.
Carefully clean the optical surface with a soft cloth until the buildup is fully removed.
3.
Clean the optics with drinking or distilled water.
4.
Clean the optics and screw on a new fluorescence cap.
NOTICE
Damages, scratches on the optical surface
Faulty measured values
► Ensure that the optical surface is not scratched or damaged in any way.
7.2
Replacing wear and tear materials
7.2.1
Replacing the sealing ring
The sealing ring must be replaced if visibly damaged. For replacement, use only original sealing
rings.
7.2.2
Replacing the fluorescence cap
The typical operating life of a fluorescence cap is more than 2 years. The sensor checks whether the
cap is aging and issues a warning via the transmitter if the rate of aging reaches a specific value. The
sensor is still able to measure at this stage. However, it is advisable to change the cap as quickly as
possible.
Removing the old fluorescence cap
1.
Remove the sensor from the medium.
2.
Unscrew the protection guard.
3.
Clean the outside of the sensor.
4.
Unscrew the fluorescence cap.
5.
Clean and dry the optical surface if necessary.
Installing the new fluorescence cap
6.
Make sure that there are no dirt particles on the sealing surface.
7.
Carefully screw the fluorescence cap onto the sensor head until the stop.
8.
Screw the protection guard back on.
After replacing the fluorescence cap, the sensor must be recalibrated. Then insert the sensor
into the medium and check that no alarm is displayed on the transmitter.
Endress+Hauser
23
Accessories
Oxymax COS61
8
Accessories
In the following sections, you find the accessories available at the time of issue of this
documentation.
For information on accessories that are not listed here, please contact your local service or sales
center.
8.1
Connection accessories
VS junction box
• With plug-in socket and 7-pole plug
• For cable extension from sensor (COS71, COS61, COS31, COS3 with SXP connector) to
transmitter, IP 65;
• Order no. 50001054
Measuring cable OMK
• for use as extension cable between junction box VS and transmitter, not terminated
• sold by the metre - order no. 50004124
COK31 special measuring cable
• for sensors COS31, COS61 and COS71 with TOP68 plug-in head
• Order numbers:
– Cable length 1.5 m (4.9 ft): 51506820
– Cable length 7 m (23 ft): 51506821
– Cable length 15 m (49 ft): 51506822
8.2
Installation accessories
Immersion assembly COA110
• for sensor immersion in the basin, PVC pipe resp. PUR floating body with SS 1.4571 (AISI 316Ti)
immersion tube
• Ordering acc. to product structure (see Technical Information TI035C/07/en)
Flow assembly COA250
• For sensor installation in pipe lines, PVC
• Ordering acc. to product structure (--> Online configurator: www.products.endress.com/
coa250)
• Technical Information TI00111C/07/EN
Retractable assembly Cleanfit COA451
• Manually driven retractable assembly, stainless steel, with ball valve, for oxygen sensors;
• Ordering acc. to product structure (--> Online configurator: www.products.endress.com/
coa451)
• Technical Information TI00368C/07/EN
Holder system Flexdip CYH112 for water
• Modular holder system for sensors and assemblies in open basins, channels and tanks
• The holder system CYH112 works for nearly any type of fixing - fixing on the floor, wall or
directly on a rail.
• Material: stainless steel
• Ordering acc. to product structure (--> Online configurator: www.products.endress.com/
cyh112)
• Technical Information TI00430C/07/EN
Wastewater assembly Flexdip CYA112
• Modular assembly system for sensors in open basins, channels and tanks
• Versions in stainless steel or PVC
• Ordering per product structure (--> Online configurator: www.products.endress.com/cya112)
• Technical Information TI00432C/07/EN
24
Endress+Hauser
Oxymax COS61
Accessories
Baffle plate OP
• extra protection for extreme flow conditions
• order no. 50028712
Membrane protection guard COY3-SK
• for sensor use in fish ponds
• order no. 50081787
8.3
Cleaning and calibration
Pressurized air cleaning system for COSXX
• Connection: 6/8 mm or 6.35 mm (¼")
• Materials: POM/V4A
• Order numbers
– 6/8 mm: 71110801
– 6.35 mm (¼"): 71110802
Compressor
• For cleaning system
• 230 V AC order number: 71072583
• 115 V AC order number: 71096199
Chemoclean
• Injector CYR10
• Ordering acc. to product structure
• Technical Information TI00046C/07/EN
Chemoclean COR3
• Spray head for sensor cleaning in immersion operation
• Material: PVC
• order no.: COR3-0
Calibration vessel
• for COS61/61D
• order no. 51518599
Endress+Hauser
25
Trouble-shooting
Oxymax COS61
9
Trouble-shooting
9.1
Trouble-shooting instructions
Problem
No display, no sensor
reaction
Check
Remedial action
Mains voltage to the transmitter?
Connect mains voltage.
Sensor connected correctly?
Set up correct connection.
Coating on the fluorescence cap?
Clean the sensor.
Medium flow available?
Create flow.
With a TOP 68 connection: humidity or dirt in Cleaning of the TOP 68 plug-in
plug?
connection by using cleaning alcohol.
Displayed value too high
Displayed value too low
Strong deviations in
displayed value
Temperature display clearly too low?
Check sensor, if necessary send
sensor in for repair.
Humidity or dirt in plug?
Cleaning by using cleaning alcohol.
Sensor calibrated?
Recalibrate
Medium flow available?
Create flow.
Displayesd temperature clearly too high?
Check sensor, if necessary send
sensor in for repair.
Coating on the fluorescence cap?
Replace
Fluorescence cap worn out?
Replace
Fluorescence cap damaged?
Replace fluorescence cap.
EMC interference on the measuring system?
Remove outer screening of sensor and
extension cable at terminal S.
Cut measuring and signalling lines from
h.v. power lines.
Make sure you comply with the instructions for troubleshooting in the Operating Instructions
of the transmitter. If necessary, carry out a test of the transmitter.
9.2
Sensor checks
Only authorised and trained personnel may test the sensor!
You will also require a multimeter (voltage, resistance).
Check
Measure
Setpoint
Voltage inspection
With the sensor connected, test the
operating voltage on the transmitter:
COM2x3-WX/WS
between terminals 87 and 0: +8 V
between terminals 88 and 0: –8 V
Slope inspection
Place the sensor in the air, and dry with a
paper towel.
After 10 minutes:
approx. 100% SAT (4 times plus-key)
Zero point inspection
Immerse the sensor in zero solution1.
Display near to 0 mg/l (0% Sat)
1
26
How to use the zero solution:
1.
Fill a large beaker (1.5 - 2 l) with approx. 1 l of water.
2.
Pour a cap-full of the zero solution into the water.
3.
Immerse the sensor into the water and wait a sufficient period of time (15 min. for oxygen
depletion).
The display drops to around 0 mg/l (0 %SAT).
Endress+Hauser
Oxymax COS61
Trouble-shooting
Depending on the conditions (contact surface water/air), the zero solution is stable for up to 12
hours.
If there are deviations from the reference values, follow the troubleshooting instructions or
contact your sales office.
9.3
Spare parts
Position
Spare parts kit
order no.
1
Sensor
acc. to product
structure
2
Sealing ring
– 2 pieces
51518597
3
Sensor cap (fluorescence cap)
51518598
1
2
without fig. Zero solution
– 3 units to produce 3 x 1 litre oxygen free solution
50001041
3
4
a0004166
Fig. 17:
9.4
Spare parts
Return
The device must be returned if repairs or a factory calibration are required, or if the wrong device
has been ordered or delivered. According to legal regulations, Endress+Hauser, as an ISO-certified
company, is required to follow certain procedures when handling returned products that are in
contact with medium.
To ensure swift, safe and professional device returns, please read the return procedures and
conditions on the internet site:
www.services.endress.com/return-material
9.5
Disposal
The device contains electronic components and must therefore be disposed of in accordance with
regulations on the disposal of electronic waste.
Please observe local regulations.
Endress+Hauser
27
Technical data
Oxymax COS61
10
Technical data
10.1
Input
Measured variable
Dissolved oxygen [mg/l, % SAT, hPa]
Temperature [˚ C, ˚ F]
Measuring range
With Liquisys COM 2x3-W:
0 to 20 mg/l (0 to 20 ppm)
0 to 200 % SAT
0 to 400 hPa
10.2
Performance characteristics
Response time
t90: 60 s
Maximum measured error
0.02 mg/l or ±1 % of measured value (< 12 mg/l)
±2 % of measured value (from 12 to 20 mg/l)
Repeatability
±0.5 % of measuring range end
Lifetime of the sensor cap
>2 years (under reference operating conditions, protect against direct sun light)
10.3
Environment
Ambient temperature
range
–20 to +60 ˚C (0 to 140 ˚F)
Storage temperature
–20 to +70 ˚C (0 to 160 ˚F)
at 95% relative humidity, non condensing
Ingress protection
• Fixed cable versions:
IP 68 (test conditions: 10 m (33 ft) water column at 25 ˚C (77 ˚F) in 30 days)
• Top 68 plug-in head versions:
IP 68 (test conditions: 1 m (3.3 ft) water column at 50 ˚C (122 ˚F) in 7 days)
10.4
Process
Process temperature
–5 to 60 ˚C (20 to 140 ˚F)
Process pressure
max. 10 bar (145 psi)
max. 10 bar (145 psi) abs.
28
Endress+Hauser
Oxymax COS61
Technical data
10.5
Mechanical construction
Weight
With cable length 7 m (23 ft): 0.7 kg (1.5 lbs.)
With cable length 15 m (49 ft): 1.1 kg (2.4 lbs.)
With TOP68 plug-in connection: 0.3 kg (0.66 lbs.)
Materials
Sensor shaft:
Cap with fluorescence layer:
Fluorescence layer:
Process connection
G1
Sensor cable
shielded 7-core fixed cable or double-shielded coaxial cable with 4 pilot wires (with TOP68 plug
connection)
Maximum cable length
max. 100 m (330 ft, (including cable extension))
Temperature
compensation
internal
Interface
RS 485
Endress+Hauser
stainless steel 1.4571 (AISI 316Ti)
POM
Silicone
29
Oxymax COS61
Index
A
I
Accessories
Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
For connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Measurement, controlling and cleaning . . . . . . . . . . . . 25
Protection gard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Ambient temperature range . . . . . . . . . . . . . . . . . . . . . . . 28
Angle of installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Application. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Ausschlagswert . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Automatic cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Basic safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Basin rim holder. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Immersion assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Immersion operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Ingress protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–8
Angle of . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Flow operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Immersion operation . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Measuring point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Retractable assembly . . . . . . . . . . . . . . . . . . . . . . . . . . 13
C
M
Cable length. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Calculating the calibration value . . . . . . . . . . . . . . . . . 19
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Chain assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Checking
Connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Configurator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Maximum cable length . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Maximum measured error . . . . . . . . . . . . . . . . . . . . . . . . 28
Measuring point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Mechanical construction . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Mounting location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
D
Occupational safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Order code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
B
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Device description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
E
Electrical connection
Direct connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Via junction box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Electromagnetic compatibility . . . . . . . . . . . . . . . . . . . . . . . 4
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Errors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
F
Field installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Floating body . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Flow operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Fluorescence cap . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18, 23
30
N
Nullpunkt. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
O
P
Panel mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Performance characteristics. . . . . . . . . . . . . . . . . . . . . . . . 28
Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Process connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Process pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Process temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Product page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Product safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Protection gard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
R
Repeatability. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Replacing
Sealing ring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Wear and tear materials. . . . . . . . . . . . . . . . . . . . . . . . 23
Requirements for personnel. . . . . . . . . . . . . . . . . . . . . . . . . 4
Response time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Retractable assembly. . . . . . . . . . . . . . . . . . . . . . . . . . 13, 24
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Endress+Hauser
Oxymax COS61
S
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Sealing ring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Sensor
Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Replacing wear and tear materials . . . . . . . . . . . . . . . . 23
Sensor cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Sensor cap. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
T
Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Mechanical construction . . . . . . . . . . . . . . . . . . . . . . . 29
Performance characteristics . . . . . . . . . . . . . . . . . . . . . 28
Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Temperature compensation . . . . . . . . . . . . . . . . . . . . . . . . 29
Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Types of calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
U
Upright post . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
W
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Wiring
Field installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Panel mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Endress+Hauser
31
BA00387C/07/EN/13.12
FM10