Servomex User and Installation Manual – Model 2700 Analyzer – 2730C Utilities Unit
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
COMBUSTION ANALYSERS
SERVOTOUGH Fluegas(2700)
2730C Utilities Unit
User and Installation Manual
Part Number:
Revision:
Language:
02730001C
1
UK English
This page intentionally blank
CONTENTS LIST
CONTENTS LIST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
LIST OF FIGURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.1
WARNINGS, CAUTIONS and NOTES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.2
Scope of this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.3
General Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.4
Unpacking. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.
DESCRIPTION AND OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2
Air Set (Single and dual sensor). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.3
Air Set and Autocalibration (Single sensor). . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.4
Air Set and Autocalibration (Dual sensor) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.5
Air Set with Blowback and Autocalibration (Single sensor) . . . . . . . . . . . . . . . . 16
2.6
Air Set with Blowback and Autocalibration (Dual sensor) . . . . . . . . . . . . . . . . . 18
2.7
Air Set and Manual Calibration (Single and dual sensor) . . . . . . . . . . . . . . . . . . 20
3.
INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.1
Mechanical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.2
Electrical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
4.
TECHNICAL SPECIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.1
Air Set (Single and dual sensor). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.2
Air Set and Manual Calibration (Single and dual sensor) . . . . . . . . . . . . . . . . . .
4.3
Air Set with Autocalibration (Single sensor) . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Air Set with Autocalibration (Dual sensor)
Air Set with Blowback and Autocalibration (Single sensor)
Air Set with Blowback and Autocalibration (Dual sensor)
29
29
30
31
5.
MAINTENANCE AND SPARES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.1
Spare parts for Single and dual sensor air set (02730701) . . . . . . . . . . . . . . . . . .
5.2
Spare parts for Single sensor air set with auto calibration (02730711) . . . . . . . .
5.3
Spare parts for dual sensor air set with auto calibration (02730713C) . . . . . . . .
5.3
Spare parts for dual sensor air set with auto calibration (02730713C) . . . . . . . .
5.4
Spare parts for Single Sensor air set with auto calibration and blowback
(02730721) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.5
Spare parts for dual sensor air set with autocalibration and blowback
(02730723C). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.6
Spare parts for single and dual sensor air set with manual calibration
(02730731) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.7
Routine maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
33
33
33
33
33
2730 User and Installation Manual
5
5
5
6
7
34
34
35
35
1
LIST OF FIGURES
Figure 1
Figure 2
Figure 3
Figure 4
Figure 5
Figure 6
Figure 7
Figure 8
Figure 9
Figure 10
Figure 11
2
Single and dual sensor air set flow diagram . . . . . . . . . . . . . . . . . . . . . . . . . .11
Single sensor air set with auto calibration flow diagram. . . . . . . . . . . . . . . . .13
Dual sensor air set with auto calibration flow diagram. . . . . . . . . . . . . . . . . .15
Dual sensor air set with blow back and auto calibration flow diagram. . . . . . .17
Dual sensor air set with blow back and auto calibration flow diagram. . . . . . .19
Single and dual sensor air set with manual calibration flow diagram. . . . . . .20
Single sensor air set installation drawing. . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Air set with either auto calibration or blowback installation drawing . . . . . .23
Gland plate details, auto calibration and blowback versions. . . . . . . . . . . . . .24
Manual calibration utility unit version installation drawing. . . . . . . . . . . . . .25
Interconnecting wiring schematic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
2730 User and Installation Manual
Revision History Sheet
02730001C 2730 Operators Manual
Ref No.
Page(s)
Affected
02730/001C/0
–
First issue
02730/001C/1
–
Filter regulator part number changed
except on assembly 02730701
2730 User and Installation Manual
Summary of Change
Date/
by
G.Botting
12.06.06
M. Hubball
4.10.10
3
This page is left intentionally blank.
4
2730 User and Installation Manual
1.
INTRODUCTION
1.1
WARNINGS, CAUTIONS and NOTES
This manual includes WARNINGS, CAUTIONS and NOTES which provide important
information relating to the following:
WARNINGS
Hazards which could result in personal injury or death.
CAUTIONS
Hazards which could result in damage to equipment or property.
NOTES
Alerts the user to pertinent facts and conditions.
1.2
Scope of this Manual
This instruction manual describes installation and operation of the 2730 Utilities Units
that are available for use with the 2700C combustion gas analyser.
•
For technical specifications, see section 4 of this manual.
•
For the addresses for technical assistance and spares, see the rear of this manual.
About this Manual
Ref.02730/001C/1
Order as Part 02730001C
CE Marking
Where applicable the 2730 utilities units carry the CE mark which indicates
conformity with the European Community Directives on CE Marking
(93/68/EEC), Electromagnetic Compatibility (EMC 89/336/EEC) and Low
Voltage Directive (LVD 73/23/EEC).
2730 User and Installation Manual
5
1.3
General Safety Information
Specific standards of safety are not included in this manual, since requirements of relevant
authorities vary widely. The safety precautions described in this manual ensure that the
existing safety features are not impaired during normal operation and use.
WARNINGS
•
•
•
•
•
•
•
6
The electrical power used in this equipment is at a voltage high enough to
endanger life. It is essential that only suitably trained and competent
personnel are allowed to access hazardous live parts by removing or
opening covers of the equipment.
The installer must be satisfied that the 2730 installation conforms to the
relevant safety requirements, National Electrical Code and any other local
regulations, and that the installation is safe for any extremes of conditions
which may be experienced in the operating environment of the equipment.
This appliance must be connected to a protective earth.
The electrical installation must include a means of isolating electrical
power by a switch or circuit breaker external to the analyser and within
easy reach of the operator. It must be marked as the disconnecting device
for the equipment.
The 2700 sensor head aspirator supply is suitable for use with
compressed air only. Flammable gases must not be used for the aspirator
supply.
The 2730 Utilities unit is not suitable for use in hazardous areas / locations
without the use of appropriate safety purge equipment.
The 2730 Utilities unit is only to be used with the gasses specified in the
2700 Quickstart Manual, 2700 Installation Manual and/or this manual.
2730 User and Installation Manual
WARNING - GAS HAZARDS
If a gas leak occurs within the automatic calibration utilities units, the
enclosure may become filled with one or both of the following gaseous
hazards.
a)
0.3% oxygen in nitrogen which is an asphyxiant.
b)
Up to 1% carbon monoxide in air which is toxic.
Appropriate care must be taken when opening the utilities enclosure for any
reason.
Automatic calibration units have a breathing device fitted to the enclosure.
Ensure that this device is not obstructed or blocked.
Manual and Automatic calibration units MUST be located in a freely ventilated
area.
1.4
Unpacking
WARNING
The 2730 Utilities Units supplied with Autocalibration weigh <16Kg (<35lb).
Care must be taken when handling.
•
Remove the 2730 components from their packing and inspect for obvious external
damage.
•
If damage has occurred, inform Servomex or its agent immediately.
•
Retain all packing and shipping information for future use.
2730 User and Installation Manual
7
This page is left intentionally blank.
8
2730 User and Installation Manual
2.
DESCRIPTION AND OPERATION
2.1
Overview
The 2700 Combustion Gas Analyser measures the concentrations of oxygen and
combustibles present in the sample, this provides the ability to maximise combustion
efficiency and reduce emissions. The 2730 Utilities Units provide a convenient package to
supply the utility gases required by the 2700 sensor head. The Utilities Unit may provide
a number of different gas supplies to the sensor head.
For optimum performance, only one analyser should be used in conjunction with an air
supply or utilities unit.
Aspirator air supply
This provides the instrument grade compressed air supply required by the
2700 combustion gas analyser sample aspirator. The aspirator is used to extract the
gas sample from the flue into the analyser sensor head. The compressed air supply
should be clean, dry, and free from oil and dirt. The aspirator pressure range
required is 3 to 5 psig. The flow rate is typically <1.5 l/min under typical flue
pressure conditions.
Auxiliary air supply
The combustibles only and dual sensor versions of the 2700 require auxiliary air
to supplement the level of oxygen present in the combustibles sensor. This is
derived from the aspirator air supply within the sensor head to give a flow of
between 80 to 120 ml/min.
Calibration gas supply
As good practice, Servomex recommend that the 2700 is checked/validated with
check gas every month and then calibrated as required. This period can be
extended as a result of operating experience. Calibration can be carried out either
automatically or manually. The utility units provide the analyser with a controlled
constant flow of the relevant calibration gases.
Blowback air supply
The blow back utilities units provide a 1 barg pressure compressed air supply to
the sensor head calibration port. In extreme conditions blowback may be used to
extend the service interval for the external sample probe filter by blowing off any
dust coating. The blow back facility is not intended as a replacement for a suitably
filtered sampling system.
2730 User and Installation Manual
9
WARNING
For the utilities units supplied with either manual or automatic calibration. Do
not unscrew the needle valve controlling the calibration gas flow rate by more
than 4 turns. Failure to do this may result in the valve assembly becoming
detached from the valve body exposing the user to potentially toxic or
asphyxiant gases.
CAUTIONS
All Utilities units are specified to -10°C (+14°F), however, if operated below
+2°C (+35.6°F) the instrument grade compressed air and calibration gases
used must be free from oil, water, dirt and have a dew point 10°C (18°F)
below the ambient temperature, to prevent freezing within the components.
Maintenance of the Utilities Units should not be performed at temperatures
below 0°C (+32°F).
The 2730 utilities unit is designed for use only with the 2700C Series
combustion gas analyser and should not be used with any other products or
systems.
NOTES
The (Hoffman/Schroff) enclosure that is supplied with the automatic
calibration utilities units requires a special key (which is provided) to gain
access for installation and service.
All gas pressures stated in this manual are ‘gauge’ pressures unless
otherwise specified.
1 barg = 100 kPag
1 barg = 14.696 psig
10
2730 User and Installation Manual
2.2
Air Set (Single and dual sensor)
Refer to the flow diagram in figure 1.
The single or dual sensor air set (part number 02730701) consists of a filtered pressure
regulator supplied complete with gauge, auto drain, inlet and outlet fittings, wall mounting
bracket, screws, washers and nuts.
The inlet pressure to the air set should be in the range 1 to 5 barg. The outlet pressure from
the regulator should be set to the required aspirator inlet pressure as stated in the sensor
head terminal box (typically 3.5 psig at a flow rate of <1.5 litres/min). The inlet and outlet
fittings are suitable for use with 1/4" OD plastic or metal tubing.
Figure 1. Single and dual sensor air set flow diagram
Aspirator Air Outlet
(3.5 psig <1.5 l/min)
typical
Air Inlet
1 to 5 barg
Auto
Drain
2730 User and Installation Manual
11
2.3
Air Set and Autocalibration (Single sensor)
Refer to the flow diagram figure 2.
The oxygen only air set with autocalibration (part number 02730711) consists of a wall
mounted enclosure inside which are fitted the following components:
•
•
•
•
•
A filtered pressure regulator supplied complete with gauge and auto drain fitting.
An unfiltered pressure regulator complete with gauge.
A multi voltage power supply.
Two solenoid valves.
A 0-1000 ml/min flow meter and fine control valve.
The inlet pressure to the unit should be in the range 1 to 5 barg. The typical flow rate
required is <2.5 litres/min. Set the outlet pressure from the filter regulator to 1 barg. The
outlet from the aspirator air precision pressure regulator is set to the required aspirator inlet
pressure as stated in the sensor head terminal box (typically 3.5 psig at a flow rate of
<1.5 litres/min). The pressure of the calibration gases to the autocalibration unit inlet port
should be set to 1 barg (flow rate of <1 litre/min). The autocalibration facility consists of
two solenoid valves to select the calibration gas required and a flow meter and needle
valve to regulate and measure the gas flow rate. The outlet flow rate from the
autocalibration unit to the calibration port of the 2700 should be set to 600 ml/min. The
inlet and outlet bulkhead fittings are suitable for use with 1/4" NPT or BSPT male fittings
and tubing.
At the 2700 control unit, to which the utilities unit is attached, configure the relay outputs
from the analyser using the ASSIGN RELAYS menu option as described in the
2700 Quickstart Manual. The relays should be set as follows.
RELAY 1
RELAY 2
12
GAS 1
GAS 2
2730 User and Installation Manual
Figure 2. Single sensor air set with auto calibration flow diagram.
Set to 1 barg
Air Inlet
1 to 5 barg
<2.5 l/min
Aspirator Air
Outlet
(3.5 psig <1.5 l/min
typical)
Air Inlet
1 to 5 barg
<2.5 l/min
SV1
Auto
Drain
SV2
0.3% Oxygen in Nitrogen -OR500ppm Carbon Monoxide in Air
Calibration Gas
Inlet
1 barg <1 l/min
2730 User and Installation Manual
Calibration Gas
Outlet
600 ml/min
13
2.4
Air Set and Autocalibration (Dual sensor)
Refer to the flow diagram in figure 3.
The dual sensor air set with autocalibration (part number 02730713C) consists of a wall
mounted enclosure inside which are fitted the following components:
•
•
•
•
•
A filtered pressure regulator supplied complete with gauge and auto drain fitting.
A precision pressure regulator complete with gauge.
A multi voltage power supply.
Three solenoid valves.
A 0-1000 ml/min flow meter and fine control valve.
The inlet pressure to the unit should be in the range 1 to 5 barg. The typical flow rate
required is <2.5 litres/min. Set the outlet pressure from the filter regulator to 1 barg. The
outlet from the aspirator air precision pressure regulator is set to the required aspirator inlet
pressure as stated in the sensor head terminal box (typically 3.5 psig at a flow rate of
<1.5 litres/min). The pressure of the calibration gases to the autocalibration unit inlet ports
should be set to 1 barg (flow rate of <1 litre/min). The autocalibration facility consists of
three solenoid valves to select the calibration gas required and a flow meter and needle
valve to regulate and measure the gas flow rate. The outlet flow rate from the
autocalibration unit to the calibration port of the 2700 should be set to 600 ml/min. The
inlet and outlet bulkhead fittings are suitable for use with 1/4" NPT or BSPT male fittings
and tubing
At the 2700 control unit, to which the utilities unit is attached, configure the relay outputs
from the analyser using the ASSIGN RELAYS menu option as described in the
2700 Quickstart Manual. The relays should be set as follows.
RELAY 1
RELAY 2
RELAY 3
14
GAS 1
GAS 2
GAS 3
2730 User and Installation Manual
Figure 3. Dual sensor air set with auto calibration flow diagram.
Set to 1 barg
Air Inlet
1 to 5 barg
<2.5 l/min
Aspirator Air
Outlet
(3.5 psig <1.5 l/min
typical)
SV1
Auto
Drain
SV2
0.3% Oxygen in Nitrogen
Calibration Gas
Inlet
1 barg <1 l/min
SV3
Calibration Gas
Outlet
600 ml/min
500ppm Carbon Monoxide in Air
2730 User and Installation Manual
15
2.5
Air Set with Blowback and Autocalibration (Single sensor)
Refer to the flow diagram in figure 4.
The single sensor air set with blowback and autocalibration (part number 02730721)
consists of a wall mounted enclosure inside which are fitted the following components:
•
•
•
•
•
A filtered pressure regulator supplied complete with gauge and auto drain fitting.
An unfiltered pressure regulator complete with gauge.
A multi voltage power supply.
Three solenoid valves.
A 0-1000 ml/min flow meter and fine control valve.
The inlet pressure to the unit should be in the range 1 to 5 barg. The typical flow rate
required is <10 litres/min. Set the outlet pressure from the filter regulator to 1 barg. The
outlet from the aspirator air precision pressure regulator is set to the required aspirator inlet
pressure as stated in the sensor head terminal box (typically 3.5 psig at a flow rate of
<1.5 litres/min). The autocalibration facility consists of two solenoid valves to select the
calibration gas required and a flow meter and needle valve to regulate and measure the gas
flow rate. The pressure of the calibration gases to the autocalibration unit inlet port should
be set to 1 barg (flow rate of <1 litre/min). The outlet flow rate from the autocalibration
unit to the calibration port of the 2700 should be set to 600 ml/min. The Blowback facility
consists of a solenoid valve and bulkhead fitting. The Outlet pressure from the Blowback
port is 1 barg. The inlet and outlet bulkhead fittings are suitable for use with 1/4" NPT or
BSPT male fittings and tubing.
At the 2700 control unit, to which the utilities unit is attached, configure the relay outputs
from the analyser using the ASSIGN RELAYS menu option as described in the
2700 Quickstart Manual. The relays should be set as follows.
RELAY 1
RELAY 2
RELAY 4
16
GAS 1
GAS 2
BLOWBACK AIR
2730 User and Installation Manual
Figure 4. Dual sensor air set with blow back and auto calibration flow diagram.
Set to 1 barg
Air Inlet
1 to 5 barg
<10 l/min
Aspirator Air
Outlet
(3.5 psig <1.5 l/min
typical)
SV4
Blowback Air
Outlet
SV1
Calibration Gas
Inlet
1 barg <1 l/min
Auto
Drain
SV2
0.3% Oxygen in Nitrogen -OR500ppm Carbon Monoxide in Air
Calibration Gas
Outlet
600 ml/min
2730 User and Installation Manual
17
2.6
Air Set with Blowback and Autocalibration (Dual sensor)
Refer to the flow diagram in figure 5.
The dual sensor air set with blowback and autocalibration (part number 02730723C)
consists of a wall mounted enclosure inside which are fitted the following components:
•
•
•
•
•
A filtered pressure regulator supplied complete with gauge and auto drain fitting.
A precision pressure regulator complete with gauge.
A multi voltage power supply.
Four solenoid valves.
A 0-1000 ml/min flow meter and fine control valve.
The inlet pressure to the unit is 1 to 5 barg. The typical flow rate required is<10 litres/min.
Set the outlet pressure from the filter regulator to 1 barg. The outlet from the aspirator air
precision pressure regulator is set to the required aspirator inlet pressure as stated in the
sensor head terminal box (typically 3.5 psig at a flow rate of <1.5 litres/min). The pressure
of the calibration gases to the autocalibration unit inlet ports should be set to 1 barg (flow
rate of 1 litre/min). The autocalibration facility consists of three solenoid valves to select
the calibration gas required and a flow meter and needle valve to regulate and measure the
gas flow rate. The outlet flow rate from the autocalibration unit to the calibration port of
the 2700 should be set to 600 ml/min. The blowback facility consists of a solenoid valve
and bulkhead fitting. The outlet pressure from the blowback port is 1 barg. The inlet and
outlet bulkhead fittings are suitable for use with 1/4" NPT or BSPT male fittings and
tubing.
At the 2700 control unit, to which the utilities unit is attached, configure the relay outputs
from the analyser using the ASSIGN RELAYS menu option as described in the
2700 Quickstart Manual. The relays should be set as follows.
RELAY 1
RELAY 2
RELAY 3
RELAY 4
18
GAS 1
GAS 2
GAS 3
BLOWBACK AIR
2730 User and Installation Manual
Figure 5. Dual sensor air set with blow back and auto calibration flow diagram.
Set to 1 barg
Air Inlet
1 to 5 barg
<10 l/min
Aspirator Air
Outlet
(3.5 psig <1.5 l/min
typical)
SV4
Blowback
Air Outlet
SV1
Auto
Drain
0.3% Oxygen in Nitrogen
Calibration Gas
Inlet
1 barg <1 l/min
SV2
SV3
Calibration Gas
Outlet
600 ml/min
500ppm Carbon Monoxide in Air
2730 User and Installation Manual
19
2.7
Air Set and Manual Calibration (Single and dual sensor)
Refer to the flow diagram in figure 6.
The single and dual sensor air set with manual calibration (part number 02730731)
consists of a wall mounted panel upon which are fitted the following components:
•
•
•
A filtered pressure regulator supplied complete with gauge and auto drain fitting.
A precision pressure regulator complete with gauge.
A 0-1000 ml/min flow meter and fine control valve.
The inlet pressure to the unit should be in the range 1 to 5 barg. The typical flow rate
required is <2.5 litres/min. Set the outlet pressure from the filter regulator to 1 barg. The
outlet from the aspirator air precision pressure regulator is set to the required aspirator inlet
pressure as stated in the sensor head terminal box (typically 3.5 psig at a flow rate of
<1.5 litres/min). The manual calibration facility consists of a flowmeter and fine flow
control valve. The pressure of the calibration gases to the manual calibration unit should
be set to 1 barg (flow rate of <1 litre/min). The outlet flow rate from the manual calibration
unit to the calibration port of the 2700 should be set to 600 ml/min. The inlet and outlet
bulkhead fittings are suitable for use with 1/4" NPT or BSPT male fittings and tubing.
Figure 6. Single and dual sensor air set with manual calibration flow diagram.
Air Inlet
1 to 5 barg
<2.5 l/min
Set to 1 barg
Aspirator Air
Outlet
(3.5 psig <1.5 l/min
typical)
Auto
Drain
Quick Connect Plug
Quick Connect Plug
20
Calibration Gas
Outlet
600 ml/min
2730 User and Installation Manual
3.
INSTALLATION
3.1
Mechanical Installation
Select a location which allows convenient access for installation and maintenance and will
minimise ambient temperature fluctuations, vibration and be freely ventilated.
The utilities units should be mounted onto a rigid vertical surface capable of supporting
the weight of the unit. The aspirator air set and manual calibration versions should be
mounted with the filter regulator drain pointing vertically down. The autocalibration
versions of the utilities units should be mounted with the electrical entries on the under
side.
Ensure that the operating ambient temperature is in the range of -10°C to +50°C and that
the Utilities Unit is protected from radiant heat sources or from direct sunlight (in ambient
temperatures over 30°C). This can be achieved by installing a Heat Shield between the
Utilities Unit and the source; a thin metal sheet is usually adequate.
All pressure regulators, flow meters and gauges exhibit variations of output with changes
in ambient temperature (their temperature coefficient). Although Servomex has chosen
products that have good rejection to ambient temperature variation, utility units should be
installed in locations with relatively stable ambient conditions. Ideally temperature swings
from one extreme to another should be avoided.
Instrument air and other gases supplied to the utilities unit must have dew points of at least
10°C less than the minimum anticipated ambient temperature.
Table 1 lists the figure numbers that contain the installation drawings for the specific
utilities units.
Table 1 – Utility Units Installation Drawings
Part No.
Description
Fig. No.
02730701
Air set (single/dual sensor)
02730711
Air set with autocalibration (single sensor)
8, 9
02730713C
Air set with autocalibration (dual sensor)
8, 9
02730721
Air set with blowback and autocalibration (single sensor)
8, 9
02730723C
Air set with blowback and autocalibration (dual sensor)
8, 9
02730731
Air and manual calibration (single and dual sensor)
10
7
NOTE
The liquid that may collect in the filter regulator bowl will be automatically
discharged through the drain port. This liquid must be disposed of in an
appropriate and safe manner.
2730 User and Installation Manual
21
Figure 7. Single sensor air set installation drawing.
f
99mm
32mm
93mm
7
8
8mm
9mm
Ø5.5mm
124mm
165mm
FLOW
1
2
3
4
5
6
Kit includes two M5 x 12mm mounting bolts, M5 washers and M5 full hex nuts
Key to figure 7.
1
2
3
4
5
6
7
8
22
Two off 5.5mm diameter mounting holes for 5mm mounting bolts.
Filter drain port (threaded 1/8" NPT male).
Filter bowl.
Outlet pressure gauge, 0-20 psig (part number 5981-2157).
Compressed air inlet (1/4" OD compression fitting).
Aspirator air outlet (1/4" OD compression fitting).
Mounting bracket.
Pressure adjustment control.
2730 User and Installation Manual
Figure 8. Air set with either auto calibration or blowback installation drawing
227mm
13
400mm
1
220mm
12
362mm
2
4
593mm
TB3
540mm
6
578mm
500mm
5
TB4
SV4
7
SV1
SV2
SV3
TB2
TB1
8
Ø10mm
9
15
10
11
Key to figure 8.
1
2
4
5
6
7
8
9
10
11
12
13
15
Hoffman hinged enclosure.
Four off 10mm diameter mounting holes.
Filter regulator assembly (part number 5984-4380).
Filter regulator output pressure gauge, 0-20 psig (part number 5981-2157).
Aspirator air pressure regulator (see spares list for part numbers).
Blowback solenoid valve (if required, part number 2372-8040).
Calibration gas solenoid valves (2 off for single sensor or 3 off for dual sensor,
part number 2372-8033).
Gland plate assembly (see figure 14).
Enclosure breather fitting (part number 2371-0076).
Power supply PCB (part number 02730901) and transparent mains protection
cover.
Calibration gas flow meter and needle valve (part number 5981-3516).
Aspirator air output pressure gauge, 0-20 psig (part number 5981-2157).
Adaptor, 1/4" BSPT (M) to 1/4" NPT (F) **.
2730 User and Installation Manual
23
Figure 9. Gland plate details, auto calibration and blowback versions.
10
DRAIN
ASP.AIR
OUTLET
1
8
9
AIR INLET
MAX. 5 BAR
CAL GAS
(O2 LOW)
MAX. 1 BAR
CAL GAS/
BLOWBACK
2
3
7
6
CAL GAS
(CO HIGH)
SIGNAL
MAX. 1 BAR
MAINS
4
5
Key to figure 9
1
2
3
4
5
6
7
8
9
10
Aspirator air outlet bulkhead fitting **.
Calibration and blowback (if fitted) outlet bulkhead fitting **.
Unused
Enclosure breather fitting (part number 2371-0076).
Main power inlet M20 cable gland.
Signal connection inlet M20 cable gland.
Combustibles sensor high calibration gas inlet bulkhead fitting **.
Oxygen sensor low calibration gas inlet bulkhead fitting **.
Compressed air inlet bulkhead fitting **.
Filter regulator automatic drain outlet **.
NOTE **
24
Gas connections are 1/4" NPT(F) as supplied, 1/4" BSPT (F) is obtained
by removal of adaptor item 15 of figure 8
2730 User and Installation Manual
Figure 10. Manual calibration utility unit version installation drawing.
16
450mm
173mm
390mm
2
15
360mm
124mm
320mm
3
4
17
14
265mm
238mm
17
41mm
5
6
Ø8.5mm
1
7
8
11
13
17
Key to figure 10
1
2
3
4
5
6
7
8
11
13
14
15
16
17
Mounting bracket.
Filter regulator assembly (part number 5984-4380).
Four off 8.5mm diameter mounting holes.
Compressed air inlet bulkhead fitting **.
6mm OD Quick Connect fitting and tubing (part number 2385-3131).
Quick connect socket.
Filter regulator output pressure gauge, 0-20 psig (part number 5981-2157).
Filter regulator automatic drain outlet.
Calibration gas flow meter and needle valve (part number 5981-3516).
Calibration gas outlet bulkhead fitting **.
Aspirator air outlet bulkhead fitting **.
Aspirator air outlet pressure gauge, 0-20 psig (part number 5981-2157).
Aspirator air pressure regulator (part number 5984-7016).
Adaptor, 1/4" BSPT (M) to 1/4" NPT (F) **.
NOTE **
Gas connections are 1/4" NPT(F) as supplied, 1/4" BSPT (F) is obtained
by removal of adaptor item 17.
- Unit comes with two 1/4" OD compression Quick connect plugs, to fit into external
calibration gasses (part number 2385-3100).
2730 User and Installation Manual
25
3.2
Electrical Installation
This section defines the electrical interconnection details for the automatic calibration
versions of the utilities units. The part numbers for the utility units requiring an electrical
power supply are 02730711, 02730713C, 02730721, and 02730723C.
WARNING
Refer to the warnings and conditions outlined in the “General Safety
Information” in section 1.3 of this manual.
CAUTIONS
•
•
•
•
The utilities unit supply voltage is field configurable for either 110 - 120V
or 220 - 240V (nominal).
Ensure that the rated voltage and frequency is suitable for the installation.
Ensure that the link on the PCB has been set to the correct supply voltage.
The supply voltage is set by linking together the terminals at terminal block
TB2. Ensure that the link setting is correct for the rated supply voltage.
The sensor head is intended to be powered from a supply which has the
‘Neutral’ conductor referenced to ground.
Where this is unavailable, it is recommended that a suitable isolating
transformer is installed and the secondary winding is referenced to
ground.
Alternatively, the electrical installation must include a fuse rated
“F 250 mA HBC” in each supply conductor. The disconnect device
(see also section 1.3) must disconnect each supply conductor. The fuses
must be marked as being associated with this equipment. Ensure that the
installation conforms to the relevant national and local safety
requirements.
NOTES
•
•
•
26
The conduit or screen (if any) associated with the power supply should be
grounded at the cable gland or conduit entry to the enclosure in
accordance with local wiring regulations.
Fuse F1 located on the PCB is rated F250mA HBC for 110-120V and
220-240V operation.
Solenoid valve 4 only must be used for the Blowback facility as it has a
larger internal diameter than Solenoid valves 1, 2 & 3.
2730 User and Installation Manual
•
•
•
•
•
•
Refer to figure 11 for an interconnection wiring schematic.
The 2730 utilities units are rated in accordance with IEC 644 for INSTALLATION
CATEGORY II which is characterised as being local level (not distribution level),
appliances and portable equipment with over voltage impulse withstand
up to 2500V.
Electrical terminations are suitable for 20 AWG (0.5 mm2) to 14 AWG (2.5 mm2)
solid conductors or 16 AWG (1.5 mm2) stranded conductors.
Gland entry sizes and positions are shown in the appropriate installation drawing.
Un-used entries should be fitted with appropriate plugs.
Ensure that good EMC practices are observed when making connections. Refer
also to the 2700 Installation manual. Where a screened cable is specified for the
relay connections, in order to avoid ground loops, the screen should be terminated
to ‘ground’ only at the Control unit.
All electrical connections and access to the fuse, F1, are made to the PCB inside
of the unit enclosure. To gain access use the special key provided and open the
door. After electrical connections are complete ensure the cover is fitted and the
door is locked closed.
Figure 11. Interconnecting wiring schematic
2700
Control Unit
2730
Control Unit
TB4
Vdc
TB4
Solenoid Valve 1
1
N/O
2
N/C
3
COM
4
N/O
5
N/C
6
COM
7
N/O
8
N/C
9
COM
10
N/O
1
RELAY 1
2
Solenoid Valve 2
3
RELAY 2
4
Solenoid Valve 3
5
RELAY 3
6
Solenoid Valve 4
11
N/C
12
COM
7
RELAY 4
8
TB3
TB1
L
Live
Nl
Neutral
E
Earth
L
Mains
Input
Mains
Input
Nl
E
0Vdc
1
Supply
2
220-240Vac
3
110-120Vac
4
100Vac
AC Voltage
Link Selector.
Shown connected
for 220-240 Vac
operation
TB2
Control unit PCB
02710901 / 027109011 / 02710921
2730 User and Installation Manual
Utilities Unit PCB
02710901
27
This page is left intentionally blank.
28
2730 User and Installation Manual
4.
TECHNICAL SPECIFICATION
4.1
Air Set (Single and dual sensor)
Part numbers - 02730701
Dimensions (W x D x H):
: 93mm x 99mm x 165mm (3.7" x 3.9" x 6.5")
Weight
: < 400 gm (< 1.0 lb)
Temperature Operating
: -10°C to +50°C (+14°F to +122°F)
Storage
: -20°C to +55°C (-4°F to +131°F)
Compressed Air Supply
Pressure
: 1 to 5 barg (15 to 72 psig)
Flowrate
: <1.5 litres/min
Instrument grade compressed air, free from oil, water and dirt.
External connections are suitable for 1/4" OD plastic or metal tubing.
Materials of Construction
Filtered Pressure Regulator
Filter bowl
Filter element
Regulator body
Adjusting knob
Elastomers
Gauge (0-20 psi / 0-1.6 bar)
Body
Face
Movement
Fittings
Wall mounting bracket
Wall mounting fixings
M5x12mm Bolts
M5 full hex nut
M5 washer
2730 User and Installation Manual
:
:
:
:
:
Polycarbonate to BS 6005
Polypropylene, (40 µm)
Zinc alloy
Acetal resin
Synthetic rubber
:
:
:
:
:
Plastic and steel
Plastic
Copper/brass
Brass (Fibre sealing washer)
Mild steel
: Stainless steel
: Stainless steel
: Stainless steel
29
4.2
Air Set and Manual Calibration (Single and dual sensor)
Part numbers - 02730731
Dimensions (W x D x H)
: 390mm x 174mm x 265mm (15.4" x 6.8" x 10.4")
Weight
: < 4.5 kg (< 10 lbs)
Temperature Operating
: -10°C to +50°C (+14°F to +122°F)
Storage
: -20°C to +55°C (-4°F to +131°F)
Compressed Air Supply
Pressure
: 1 to 5 barg (15 to 72 psig)
Flow rate
: <2.5 litres/min
Instrument grade compressed air, free from oil, water and dirt.
Calibration Gas Supply
Pressure
: 1 barg (15 psig)
Flow rate
: 1 litre/min
Customer connections are brass /nickel plated brass and suitable for use with 1/4" NPT or
BSPT male fittings.
Materials of Construction
Filtered Pressure Regulator
Filter bowl
Filter element
Regulator body
Adjusting knob
Elastomers
Gauge (0-20 psi / 0-1.6 bar)
Body
Face
Movement
Precision Pressure Regulator
Regulator body
Adjusting knob
Elastomers
Flowmeter & Valve
O rings
Calibration Quick fit connections
Body
Seals
Wall mounting bracket
Bulkhead /Adaptor fittings
Fittings
Tubing
30
:
:
:
:
:
Polycarbonate to BS 6005
Polypropylene, (40 µm)
Zinc alloy
Acetal resin
Synthetic rubber
: Plastic and steel
: Plastic
: Copper/brass
: Zinc alloy
: Acetal resin
: Synthetic rubber
: Polycarbonate and steel
: Neoprene and Buna N
:
:
:
:
:
:
Stainless steel and brass
Buna
Epoxy powder painted mild steel
Brass /nickel plated brass
Plastic & Steel with NBR packing & ‘O’ rings
Polyurethane
2730 User and Installation Manual
4.3
Air Set with Autocalibration (Single sensor)
Air Set with Autocalibration (Dual sensor)
Air Set with Blowback and Autocalibration (Single sensor)
Air Set with Blowback and Autocalibration (Dual sensor)
Part numbers - 02730711, 02730713C, 02730721 & 02730723C
Dimensions (W x D x H)
: 400mm x 220mm x 500mm (16" x 9" x 20")
Wall mounting brackets
: add an extra 40mm top and bottom
Weight
: < 16 kg (< 35 lb)
Ingress protection
: IP65/NEMA 12
Power Supply
: 110 - 120 Vac, 50/60 Hz, <20VA
(Field configurable)
: 220 - 240 Vac, 50/60 Hz, <20VA
Installation Category
: II
Temperature Operating
: -10°C to +50°C (+14°F to +122°F)
Storage
: -20°C to +55°C (-4°F to +131°F)
Compressed Air Supply
Pressure
: 1 to 5 barg (15 to 72 psig)
Flow rate
: <2.5 litres/min without blowback
: <10 litres/min with blowback
Instrument grade compressed air, free from oil, water and dirt.
Calibration Gas Supply
Pressure
: 1 barg (15 psig)
Flow rate
: 1 litre/min
Customer connections are nickel plated brass and suitable for use with 1/4" NPT or BSPT
male fittings.
Materials of Construction
Filtered Pressure Regulator
Filter bowl
Filter element
Regulator body
Adjusting knob
Elastomers
Gauge (0-20 psi / 0-1.6 bar)
Body
Face
Movement
Precision Pressure Regulator
Regulator body
Adjusting knob
Elastomers
Solenoid Valve
Body
Seal
Flowmeter & Valve
O rings
Enclosure
Wall mounting brackets
Bulkhead / Adaptor fittings
Internal Fittings
2730 User and Installation Manual
:
:
:
:
:
Polycarbonate to BS 6005
Polypropylene, (40 µm)
Zinc alloy
Acetal resin
Synthetic rubber
: Plastic and steel
: Plastic
: Copper/brass
: Zinc alloy
: Acetal resin
: Synthetic rubber
:
:
:
:
:
:
:
:
Stainless steel
Viton
Polycarbonate and steel
Neoprene and Buna N
Powder coated steel (RAL7032)
Zinc plated steel
Nickel plated brass and plastic
Plastic & Steel with NBR packing & ‘O’ rings
31
Tubing
Venting device
: Polyurethane
: Stainless steel
The four automatic calibration utilities units 02730711, 02730713C, 02730721, &
02730723C comply with the “CE Marking Directive” 93/68/EEC.
32
2730 User and Installation Manual
5.
MAINTENANCE AND SPARES
5.1
Spare parts for Single and dual sensor air set (02730701)
5.2
5.3
Part No.
Description
5984-4372
Filter regulator <3.5 bar
5981-2157
Pressure gauge 0-20 psig
Spare parts for Single sensor air set with auto calibration (02730711)
Part No.
Description
5984-4380
Filter regulator <7 bar
5984-4295
Aspirator air pressure regulator
5981-2157
Pressure gauge 0-20 psig
2372-8033
Calibration gas solenoid valve
02730901
Power supply PCB
2371-0076
Enclosure breather fitting
5981-3516
Dwyer flow meter 0.2 - 1.0 l/min
Spare parts for dual sensor air set with auto calibration (02730713C)
Part No.
Description
5984-4380
Filter regulator <7 bar
5981-2157
Pressure gauge 0-20 psig
5984-7016
Precision pressure controller
2372-8033
Calibration gas solenoid valve
02730901
Power supply PCB
2371-0076
Enclosure breather fitting
5981-3516
Dwyer flow meter 0.2 - 1.0 l/min
2730 User and Installation Manual
33
5.4
5.5
34
Spare parts for Single Sensor air set with auto calibration and blowback
(02730721)
Part No.
Description
5984-4380
Filter regulator <7 bar
5984-4295
Aspirator air pressure regulator
5981-2157
Pressure gauge 0-20 psig
2372-8033
Calibration gas solenoid valve
2372-8040
Blowback gas solenoid valve
02730901
Power supply PCB
2371-0076
Enclosure breather fitting
5981-3516
Dwyer flow meter 0.2 - 1.0 l/min
Spare parts for dual sensor air set with autocalibration and blowback
(02730723C)
Part No.
Description
5984-4380
Filter regulator <7 bar
5981-2157
Pressure gauge 0-20 psig
5984-7016
Precision pressure controller
2372-8033
Calibration gas solenoid valve
2372-8040
Blowback gas solenoid valve
02730901
Power supply PCB
2371-0076
Enclosure breather fitting
5981-3516
Dwyer flow meter 0.2 - 1.0 l/min
2730 User and Installation Manual
5.6
5.7
Spare parts for single and dual sensor air set with manual calibration
(02730731)
Part No.
Description
5984-4380
Filter regulator <7 bar
5981-2157
Pressure gauge 0-20 psig
5984-7016
Precision pressure controller
5981-3516
Dwyer flow meter 0.2 - 1.0 l/min
2385-3100
Quick connect plug (1/4" OD compression)
2385-3131
Quick connect plug (6mm OD compression)
Routine maintenance
CAUTIONS
All Utilities units are specified to -10°C (+14°F), however, if operated below
+2°C (+35.6°F) the instrument grade compressed air and calibration gases
used must be free from oil, water, dirt and have a dewpoint 10°C (18°F)
below the ambient temperature, to prevent freezing within the components.
Maintenance of the Utilities Units should not be performed at temperatures
below 0°C (+32°F).
•
On initial installation check the filter regulator bowl monthly for signs of
contamination or liquid from the compressed air inlet stream. If required remove
and clean the bowl and filter. If no contamination is evident then this time interval
can be extended depending on experience.
•
Every 3 months visually inspect the internal and external pipe work and fittings for
signs of leaks or failure.
•
Every 6 months check and if necessary adjust the aspirator air pressure and
calibration gas flow rates.
•
Autocalibration units in enclosures - ensure that the breather is not obstructed.
2730 User and Installation Manual
35
NOTES
36
2730 User and Installation Manual