Yokogawa Model TB750G Right Angle Scattered Light Turbidimeter Quick Start Manual
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
User’s Manual Model TB750G Right Angle Scattered Light Turbidimeter Quick Start Manual IM 12E01A06-02E R IM 12E01A06-02E 2nd Edition i INTRODUCTION n Structure of this Manual This is a supplementary manual to TB750G Right Angle Scattered Light Turbidimeter for User's Manual IM 12E01A06-01E. This manual describes the specifications, installation, operation, maintenance, and troubleshooting. To use this instrument correctly, read this manual and User's Manual IM 12E01A0601E thoroughly. n Before Measurement The TB750G turbidimeter is preset with defaults prior factory shipment. Before measurement, verify that these factory default settings meet the operating conditions and if necessary, reconfigure parameters.  Symbol and Notations Used in this Manual n Symbol The following symbol is used in this manual. CAUTION Indicates that damage to software or hardware, or system failures could occur if instructions are not followed. n Notations In this manual, operation keys, display messages, indicators, and LED lamps on the instrument are denoted in the following manner. 1. Operation keys are denoted by brackets [ ] – for example: YES : [YES] key 2. Indicators, messages, and numerical data on the display are denoted as follows: Message display: {*WASH} Data display: {2.05} (when lit), {2.05} (when flashing) Media No.IM 12E01A06-02E (CD) 2nd Edition :Mar. 2015 (YK) All Rights Reserved Copyright © 2006, Yokogawa Electric Corporation IM 12E01A06-02E ii  Notice n About this Manual l This manual should be passed on to the end user. l This manual should be read thoroughly before operating the instrument. l This manual explains the functions contained in this product, but does not warrant that they will suit the particular purpose of the user. l The contents of this manual shall not be reproduced or copied, in part or in whole, without permission. l The contents of this manual are subject to change without prior notice. l Every effort has been made to ensure the accuracy in the preparation of this manual. However, if any errors or omissions are noticed, please contact the nearest Yokogawa representative or sales office. n Protection, Safety, and Modification of the Product l The safety instructions described in this manual should be strictly observed to ensure safety both of the product and the system controlled by the product. l A protection or safety circuit should be installed externally, if needed. Do not attempt to modify or add such circuit to the inside of the equipment. n Limitation of Liability l Yokogawa grants no warranties other than the express warranty set forth under the warranty provisions. l Yokogawa shall not be liable to you or any third party for any damage, including consequential or incidental damages, arising out of or in connection with the use of this equipment, defects beyond our knowledge, or any other contingency beyond our control. IM 12E01A06-02E Toc-1 Model TB750G Right Angle Scattered Light Turbidimeter Quick Start Manual IM 12E01A06-02E 1st Edition CONTENTS INTRODUCTION........................................................................................................i  Symbol and Notations Used in this Manual.................................................................. i  Notice................................................................................................................................ ii 1. OVERVIEW................................................................................................. 1-1 1.1 Features.............................................................................................................. 1-1 1.2 2. Measurement Principle..................................................................................... 1-1 PIPING AND WIRING................................................................................. 2-1 2.1 Piping.................................................................................................................. 2-1 2.1.1 System Using Open Head Tank and Zero Turbidity Filter.................. 2-1 2.1.2 System without Head Tank and Zero Turbidity Filter.......................... 2-2 2.2 Wiring.................................................................................................................. 2-3 2.2.1 Wiring Required for TB750G.............................................................. 2-3 2.2.2 Wiring for Detector.............................................................................. 2-4 2.2.3 Wiring for Converter............................................................................ 2-5 3. OPERATION............................................................................................... 3-1 3.1 Preparation......................................................................................................... 3-1 3.1.1 Outline................................................................................................. 3-1 3.2 3.1.2 Converter Operation Panel................................................................. 3-1 3.1.3 Calibration........................................................................................... 3-2 3.1.4 Performing Zero/span Calibration Using Filtered Water as Zero Reference................................................................................... 3-3 3.1.5 Performing 2-point Calibration Using Standard Solution.................... 3-7 Normal Operation............................................................................................... 3-8 3.2.1 When Sample Water Supply is Cut Off............................................... 3-8 3.2.2 Shutdown............................................................................................ 3-8 3.2.3 Restart................................................................................................. 3-8 4. PARAMETER SETTING............................................................................. 4-1 4.1 Outline of Operation, Setting, and Service Levels......................................... 4-1 4.2 Operation Level.................................................................................................. 4-2 4.3 Setting Level....................................................................................................... 4-3 4.4 Service Level...................................................................................................... 4-4 4.5 Procedures for Setting Parameters................................................................. 4-6 IM 12E01A06-02E Toc-2 5. MAINTENANCE.......................................................................................... 5-1 5.1 Lamp Replacement............................................................................................ 5-1 5.2 Calibration.......................................................................................................... 5-2 5.2.1 Outline................................................................................................. 5-2 5.2.2 2-point Calibration Using Turbidity Standard Solutions...................... 5-2 6. TROUBLESHOOTING............................................................................... 6-1 Revision Information................................................................................................i IM 12E01A06-02E 1-1 <1.Overview> 1. OVERVIEW 1.1 Features The TB750G is a process turbidimeter employing the right angle light scattering method, and has the following features. • Highly reliable measurement with excellent linearity and repeatability - Linearity: ±2% of reading or ±0.01 NTU, whichever is greater - Repeatability: ±1% of reading or ±0.002 NTU, whichever is greater - Display resolution: 0.001 NTU • Easy-to-clean measurement cell • Compact, lightweight converter and detector • User configurable analog output range - Analog output range: 0-0.2 NTU to 0-100 NTU • Analog output range switching (2 or 3 ranges) • Enhanced self-diagnostic function as standard - Light source failure, input element failure, calibration failure, various circuit failures, etc. • Detector designed to remove influence of air bubbles • A wide range of measurement conditions - Low flow rate: 0.05 to 20 l/min {0.8 to 317 gal/h} - High pressure: Up to 500 kPa {72 psi} - Sample temperature: 0 to 50°C {32 to 122°F} • Detector can be connected for in-line analysis • 2 analog outputs, 3 relay contact outputs, and 1 serial communication • Many options available upon request • Ultrasonic transducer and oscillator for ultrasonic cleaning • Various head tanks to accommodate application requirements 1.2 Measurement Principle The measuring system of the TB750G turbidimeter, comprised of a detector and a converter, complys with USEPA 180.1. Scatted Light Detection Element Scatted Light Detector Receiver (Darkroom) Light Source Lens (Lamp) Measuring Water Measurement Window (Glass) F1-1E.ai Figure 1.1 Measurement Principle IM 12E01A06-02E Blank Page 2. 2-1 <2. PIPING AND WIRING> PIPING AND WIRING 2.1 Piping Piping methods depend on the application, select the appropriate system configuration, and install pipes accordingly following examples. 2.1.1 System Using Open Head Tank and Zero Turbidity Filter This is a typical system. A piping diagram is shown in Figure 2.1. Air bubbles in a water sample are removed by an open head tank and the water sample is introduced into the detector at a stable flow rate. Head Tank H Flow Control Valve Sample Water Outlet Sample Water Drain Valve Valve Drain Tap Water Drain TB750G Detector h Sample Water Sample Water Effluent Drain Sample Water Sample Water Drain Port Inlet Sample Water Drain Valve Supply Valve Zero Water Drain Zero Water Supply Valve Tap Water Valve Zero Water Drain Valve Zero Turbidity Filter Other Turbidity Detectors F2-1E.ai Figure 2.1 Piping Diagram • To prevent corrosion of water inside the zero turbidity filter, allow water to flow through the filter and flow out from the zero water drain valve continuously at a flow rate of approximately 10 ml/ min. • When using filtered water as zero reference, install a filter with the appropriate pore size depending on the measuring range. Measuring range of less than 2 NTU: 0.2 µm Measuring range of 2 NTU or greater: 1 µm IM 12E01A06-02E 2-2 <2. PIPING AND WIRING> 2.1.2 System without Head Tank and Zero Turbidity Filter This is a simple system where a water sample is taken from the process and directly introduced into the detector. This system configuration can be used when a water sample contains a negligible amount of air bubbles or when the turbidity of a water sample is high and the effect of air bubbles is nonsignificant. CAUTION Sample water conditions and ambient temperature must meet the specifications: maximum pressure: 500 kPa; sample temperature: 0 to 50°C; ambient temperature: -5 to 50°C. Flow Control Valve Sample Water Effluent Sample Water Outlet Sample Water Sample Water Sample Water Valve Tap Water Zero Water Sample Water Inlet Sample Water Supply Valve h TB750G Detector Process Piping Drain Drain Port Drain Valve Drain Zero Water Supply Valve Tap Water Valve Zero Water Drain Valve Zero Turbidity Filter Other Turbidity Detectors F2-2E.ai Figure 2.2 Piping Diagram • The flow control should not be done on the inlet side of the detector. • To prevent corrosion inside the zero turbidity filter, allow water to flow through the filter and flow out from the zero water drain valve continuously at a flow rate of approximately 10 ml/min. IM 12E01A06-02E 2-3 <2. PIPING AND WIRING> 2.2 Wiring 2.2.1 Wiring Required for TB750G TB750G Turbidity detector TB750G Turbidity converter L CONVERTN POWER ER G G Power supply *5 Grounding *1 (100Ω or less) L N TUS G Dedicated cable (1/2/3m) U1U2 Grounding *1 (100Ω or less) Power supply cable *6 TUS400G Ultrasonic Oscillator *5 L2 L1 SENSOR S1 NO NC COM Contact output S1 S2 NO NC COM Contact output S2 FAIL NO NC COM Contact output FAIL RANGE A B C COM Range contact output G Serial communication *3 *4 (RS-422) Ultrasonic transducer connecting cable (maximum length: 15m) (customer wiring) Contact input U1 U2 U3 S Analog output 1 (4-20mA DC) *3 Analog output 2 (4-20mA DC or 0-20mA DC) *3 Grounding *2 (100Ω or less) RX+ RXTX+ RS-422 TXSG G IN1 IN2 COM + + G INPUT mA1 mA2 OUTPUT Grounding *1 (100Ω or less) (Note) Dotted wiring is external wiring. Use cable with 6 to 12 mm OD for wiring. *1 Power terminal "G" on detector, detector case, and converter case must be grounded (ground resistance: 100Ω or less). *2 External grounding terminal of ultrasonic oscillator must be grounded (ground resistance: 100Ω or less). *3 Use 2-conductor shielded cable for analog output wiring and serial communication wiring. *4 The wiring configuration is described below in case that RS-232C serial communication is selected. Serial communication (RS-232C) *5 *6 TXD RXD RS-232C SG G When option code "/US" is specified, TUS400G should be purchased separately. When TUS400G is used in system, the power supply to TB750G should be the same as the supply voltage specified in the MS Code of TUS400G. When "TB750G/US" and "TUS400G-NN-RC" or "TUS400G-NN-KC" is specified, refer to IM 19C1B4-01E for wiring. Figure 2.3 F2-3E.ai Connection Diagram IM 12E01A06-02E 2-4 <2. PIPING AND WIRING> 2.2.2 Wiring for Detector Wire for connections to the following terminals: (1) Power supply terminals; (2) Power supply terminals for ultrasonic oscillator (when option code “/US” is specified); and, (3) External grounding terminal. Terminals TM3 (Power to Lamp) (1) Power Supply Terminals (2) Power Supply Terminals for Ultrasonic Oscillator Figure 2.4 Terminal G (For Converter - Detector Connecting Cable) Connector for Converter - Detector Cable F2-4E.ai Internal View of Detector (3) External Grounding Terminal (M5 Screw) (Norminal size of at least 2 mm2) F2-5E.ai Figure 2.5 IM 12E01A06-02E Grounding Terminal of Detector 2-5 <2. PIPING AND WIRING> 2.2.3 Wiring for Converter Wire for connections to the following terminals: (1) Analog output/contact input terminals; (2) Contact output/range contact output terminals; (3) Serial communication terminals (when in use); and, (4) Grounding terminal. (A) Converter - Detector Connecting Cable Front Cover Serial Communication Terminals Connector for Converter - Detector Cable (3) Serial Communication Front Cover Terminals Terminal Cover (B) Terminal G (For Converter Detector Connecting Cable) (1) Analog Output/Contact Input Terminals (2) Contact Output/ Range Contact Output Terminals Analog Output/Contact Input Terminals F2-6E.ai How to Remove the Terminal Cover Disconnect the converter-detector connecting cable (A) and its grounding wire (B) from the converter. Remove the terminal cover by sliding it slightly to the left and pulling it forward. Figure 2.6 Internal View of Converter Toothed Lock Washer (4) Grounding Terminal (M4 Screw) Grounding Wire (Nominal Size of at least 2 mm2) F2-7E.ai Figure 2.7 Grounding Terminal of Converter IM 12E01A06-02E 2-6 <2. PIPING AND WIRING> Contact Output (S1, S2 and FAIL) Wiring • Be sure to use the contacts meeting the conditions below. • Functions of contact outputs S1 and S2 should be set in Codes 40 and 41 at service level. Refer to section 4.3. Table 2.1 Contact Rating AC DC Contact maximum permissible voltage 250 V 30 V Contact maximum permissible current 2A 3A Contact maximum permissible power (resistance load) 125 VA 60 W Table 2.2 Contact Operation Contact S1, S2 Status LED In action ON Not in action OFF Open Closed OFF Closed Open Power OFF OFF Open Closed OFF IM 12E01A06-02E NO NC Contact FAIL Closed Open LED ON NO NC Open Closed Open Closed <3. OPERATION> 3-1 3. OPERATION 3.1 Preparation 3.1.1 Outline • Make sure that TB750G converter and detector are fixed fimly. • Unused cable inlets should be plugged to prevent moisture from entering. 1. Installation Check 2. Piping Check 3. Wiring Check • Note that the TB750G does not have an internal power switch. 4. Supplying Power • Upon power up, allow the meter to warm up for at least one hour. • When using zero turbidity filter, allow water to run through it for at least one hour. 5. Warm-up 6. Setting and Checking Parameters • For details of parameter settings, refer to chapter 4. 7. Calibration • For details of calibrations, refer to the following pages. 8. Supplying a Water Sample and adjusting the flow rate 9. Operation check • Adjust the opening of the valve at the sample water out of the detector. • Make sure that the system is free of defects. F3-1E.ai 3.1.2 Converter Operation Panel Turbidity Unit Indication Operation Key ([*]: To enter the operation level from measurement mode) Contact Status LED LCD Screen Operation Key F3-2E.ai IM 12E01A06-02E 3-2 <3. OPERATION> 3.1.3 Calibration (1) When performing a calibration using 0.2 or 1.0 µm filtered water as zero reference, refer to section 3.1.4. (2) When performing a 2-point calibration using standard solutions, refer to section 3.1.5. Use this method to perform a calibration complying with EPA 180.1. Note: The instrument has been pre-calibrated using 0.2 µm filtered water as zero reference at the factory before shipment. Calibration Flow Chart Start calibration NO Is filtered water used as zero reference? YES Perform span/zero calibration using filtered water Perform 2-point calibration using standard solutions* Perform zero calibration Perform point calibration first using zero solution Is span calibration performed using check block or standard solution? Standard Solution Check block Perform span calibration using check block. Refer to Section 3.1.4. Perform span calibration using standard solution. Refer to Section 3.1.4. Perform second point calibration using span solution. Refer to Section 3.1.5. End calibration * Calibration complying with EPA Method 180.1. IM 12E01A06-02E F3-3E.ai 3-3 <3. OPERATION> 3.1.4 Performing Zero/span Calibration Using Filtered Water as Zero Reference START • Supply power • Warm-up (at least 1 hour) • Allow water to run through the zero filter • Supply zero water • Prepare span solution, if necessary [MODE] CALB [YES] X.XXX STD.CAL Automatically goes into maintenance mode and holds the output. [YES] X.XXX ZERO • Confirm reading has stabilized (3 to 5 min) [YES] 0.000 VALUE {0.000} None configurable [ENT] X.XXX ZERO X.XXX CAL.END Auto-adjustment Zero calibration is complete [NO] A To span calibration using check block [YES] B To span calibration using standard solution F3-4E.ai IM 12E01A06-02E 3-4 <3. OPERATION> Span calibration using check block Note: For span calibration using standard solution, refer to next page. From zero calibration A X.XXX STD.CAL [YES] X.XXX ZERO [NO] X.XXX SPAN • Stop supplying zero water • Drain zero water • Install check block * • Supply zero water • Confirm reading has stabilized (3-5 min) [YES] XX.XX VALUE {XX.XX} None configurable • Check that display reads the value of check block [ENT] XX.XX SPAN XX.XX CAL.END Auto-adjustment Span calibration is complete [YES] HOLD Release HOLD ** [NO] END * How to install check block 1. Remove the rubber cover from the top of the detector. 2. Remove the retaining plate on the window of the top cover by loosening the 2 setscrews. 3. Detach the observation window and replace it with the check block. Fix it with the 2 setscrews securely. Check block ** Be sure to release HOLD IM 12E01A06-02E F3-5E.ai <3. OPERATION> 3-5 Span calibration using standard solution (1/2) From zero calibration Before performing span calibration using standard solution, standard solution/check block selection should be set to "0 (standard solution)" in CODE 16 as shown below. B HOLD [YES] HOLD [*] at least second To enter operation level *SETP [NO] [NO] [NO] (3 times) {*SETP} -> {*RANGE} -> {*HOLD} -> {*SERV} *SERV [YES] 16 *CODE [] [] • Enter "16" using []and [] keys [YES] 0 *CAL.TP [] • Enter "0" using [] key [ENT] *SERV [MODE] HOLD [YES] HOLD [MODE] C F3-6E.ai IM 12E01A06-02E 3-6 <3. OPERATION> Span calibration using standard solution (2/2) C CALB [YES] X.XXX STD.CAL [YES] X.XXX ZERO [NO] X.XXX SPAN • Remove top cover (6 setscrews) • Rinse the cell with span solution and drain • Fill the cell with span solution to 90% * • Replace top cover with setscrews • Confirm reading has stabilized (3-5 min) [YES] 00.00 VALUE [] • Set decimal point using [] key [ENT] 20.00 ** VALUE [] [] • Enter value of span calibration using [] and [] keys [ENT] XX.XX SPAN XX.XX CAL.END Auto-adjustment Span calibration is complete [YES] HOLD Release HOLD *** [NO] END * How to pour span solution Pour span solution into the measurement cell very slowly, using a funnel or relevant equipment, taking care not to allow air bubbles to enter the cell. ** Number in the figure is an example. *** Be sure to release HOLD IM 12E01A06-02E F3-7E.ai <3. OPERATION> 3-7 3.1.5 Performing 2-point Calibration Using Standard Solution START • Supply power • Warm-up (at least 1 hour) • Wash the cell • Prepare zero and span solutions [MODE] CALB [YES] X.XXX STD.CAL [NO] X.XXX SMP.CAL [YES] X.XXX ZRSPAN • • • • • Remove top cover (6 setscrews) Rinse the cell with zero solution and drain Fill the cell with zero solution to 90% * Replace top cover with screws Confirm reading has stabilized (3-5 min) [YES] 0.000 VALUE.Z [] • Set decimal point using [] key [ENT] 0.020 ** VALUE.Z [] [] • Enter value of zero calibration using [] and [] keys [ENT] X.XXX ZERO Auto-adjustment X.XXX SPAN • Remove top cover (6 setscrews) • Rinse the cell with zero solution and drain • Fill the cell with zero solution to 90% * • Replace top cover with screws • Confirm reading has stabilized (3-5 min) [YES] 00.00 VALUE.S [] • Set decimal point using [] key [ENT] 20.00 ** VALUE.S [] [] • Enter value of span calibration using [] and [] keys [ENT] XX.XX SPAN XX.XX CAL.END Auto-adjustment Span calibration is complete [YES] HOLD Release HOLD *** [NO] END * How to pour zero solution Pour a span solution into the measurement cell very slowly, using a funnel or relevant equipment, taking care not to allow air bubbles to enter the cell. ** Numbers in the figures are examples. *** Be sure to release HOLD F3-8E.ai IM 12E01A06-02E 3-8 <3. OPERATION> 3.2 Normal Operation The TB750G turbidimeter does not normally require any manual operations except when periodic cleaning/calibration is performed or when a failure occurs. 3.2.1 When Sample Water Supply is Cut Off The turbidity measurement is not affected by temporary suspension of sample water supply, but longterm suspension may interfere with the accuracy of the instrument. Since the TB750G cannot detect whether the sample water supply is stopped or not, periodical check of the sample line is needed. 3.2.2 Shutdown The parameter settings and other information set in the turbidimeter are retained even if power is turned off. If the instrument will be out of operation for a long period, power should be removed. Dirt and/or stains on the instrument should be washed off thoroughly when the turbidimeter is to be removed from the site. Rinse the inside of the measurement cell, and then empty the cell or keep zero water running through the cell. 3.2.3 Restart When power is returned, the turbidimeter is put into the measurement state. It requires at least one hour for warm-up. Check that the reading stabilizes well, before performing calibration. IM 12E01A06-02E 4-1 <4. PARAMETER SETTING> 4. PARAMETER SETTING 4.1 Outline of Operation, Setting, and Service Levels Parameters should be set in the relevant mode classified into three levels: operation, setting and service. Measurement Mode [MODE] key Operation Level [ * ] For at least second Setting Level CALIB (Calibration) STEP (Alarm setting) DISP (Message display) RANGE (Analog output) STEP (Alarm setting) HOLD (Hold selection) HOLD (Hold selection) Service Level CODE XX Figure 4.1 F4-1E.ai Transition between Measurement Mode and Three Levels IM 12E01A06-02E 4-2 <4. PARAMETER SETTING> 4.2 Operation Level This is the level where, basically, operations regarding routine inspections and maintenance, such as calibration and manual cleaning, can be done. Also, the desired display item on the message display can be selected at this operation level. Measurement mode [MODE] CALIB [YES] [YES] Calibration mode HOLD [NO] DISP [YES] [YES] Message display selection mode [NO] *1 SETP [YES] [YES] Alarm setting mode [NO] *2 HOLD [YES] HOLD *3 [YES] Hold selection mode [NO] *1: Skipped if alarm setting function is set to “0: Disabled” in Code 51 at service level, or unless either of functions of contact output S1 or S2 is set to “1: Alarm” in Code 40 or 41 at service level. *2: Skipped if hold function is set to “*H.OFF: Disabled” in SET HOLD mode at setting level. *3: If [YES] key is pressed, analog output is held in measurement mode. F4-2E.ai Table 4.1 Setting Parameters at Operation Level Mode CAL Setting Parameter Calibration Using turbidity standards, or using water sample Start Turbidity value Self-adjustment in process End Display Default STD.CAL or SMP.CAL ZERO or SPAN or ZRSPN VALUE 0.001 to 2000[NTU] *1 ZERO or SPAN CAL.END CAL START Not available WASH START DISPLAY Setting Range CALIB Not available Message display selection Analog output 1 current Analog output 2 current Switchable output range state DISP XX.XmA1 Output 1 current XX.XmA2 RNG1/2-A/B/C *2 * SETP *3 (No pointer at Alarm setting mode display) High alarm setpoint * S.TB-H –010.0 to 2200.0[NTU] *1 2200.0(NTU) Low alarm setpoint * S.TB-L –010.0 to 2200.0[NTU] *1 -010.0(NTU) HOLD Hold selection Hold selection *1: *2: *3: *4: IM 12E01A06-02E HOLD *4 (Disabled) Although setting range is up to 2000 NTU, measuring range is up to 100 NTU. Skipped if analog output range switching selection {*RNGPR} is set to “0: Fixed” in CODE 30 at service level. Skipped if alarm setting function is set to “0: Disabled” in Code 51 at service level, or if either function of contact output S1 or S2 is not set to “1: Alarm” in Code 40 or 41 at service level. Skipped if hold function is set to “*H.OFF: Disabled” in SET HOLD mode at setting level. 4-3 <4. PARAMETER SETTING> 4.3 Setting Level There are modes where parameters regarding output signals and contact outputs can be set. *2 Measurement mode HOLD [ * ] For at least second *1 *SETP [YES] SETPOINTS Mode [NO] *RANGE [YES] RANGE Mode [NO] *HOLD [NO] *SERV [YES] *3 [YES] SET HOLD Mode Service Level [NO] *1: Skipped unless either of functions of contact output S1 or S2 is set to “1: Alarm” in Code 40 or 41 at service level. *2: Skipped if hold function is set to “*H.OFF: Disabled” in SET HOLD mode at setting level. *3: For service level, see Section 4.4. Table 4.2 Mode Setting Parameters at Setting Level Setting Parameter SETPOINTS Alarm setting RANGE Display * SETP Setting Range *1 High alarm setpoint * S.TB-H –010.0 to 2200.0[NTU] *1 2200.0(NTU) * S.TB-L –010.0 to 2200.0[NTU] *1 -010.0(NTU) Analog output range setting * RANGE Zero point Span point Local range selection Local range for output 1, or local range for output 2 Hold parameter setting * FIXR.1 or *FIXR.2 * ZERO 0000.00 to 2000.00[NTU] 0000.00(NTU) * SPAN 0000.00 to 2000.00[NTU] 0100.00(NTU) * LOCAL *LCL.1 or *LCL.2 0: Range A, 1: Range B, 2: Range C 0:Range A * HOLD Hold function enabled, or hold function disabled * H.ON or *H.OFF Last measured value, or fixed value * H.LST or * H.FIX *H.ON (Enabled) *H.OFF (Disabled) * H.ON:Enabled *H.LST (Last measured value) *H.FIX (Fixed value) * H.LST:Last measured value Fixed value for output 1 * H.mA1 02.0 to 22.0[mA] 22.0(mA) Fixed value for output 2 * H.mA2 00.0 to 22.0[mA](0-20mA) 02.0 to 22.0[mA](4-20mA) 22.0(mA) CAL/WASH Not available SERVICE *1: *2: Default Low alarm setpoint Fixed range for output 1, or fixed range for output 2 SET HOLD F4-3E.ai *2 Skipped if alarm setting function is set to “0: Disabled” in Code 51 at service level, or if either function of contact output S1 or S2 is not set to “1: alarm” in Code 40 or 41 at service level. For service level, see Section 4.4. IM 12E01A06-02E 4-4 <4. PARAMETER SETTING> 4.4 Service Level The TB750G turbidimeter has various functions. At this level there are modes where functions can be selected according to need for the operation. Table 4.3 Code CODE02 CODE08 CODE11 CODE12 CODE13 CODE14 CODE16 CODE17 CODE30 CODE33 Setting Parameters at Service Level (1/2) Setting Parameter Input voltage display Transmitted light detection element input display Scattered light detection element input display End Spike detection parameter setting Spike detection function enabled or disabled Limit value Hold time Sampling time Zero calibration factor display Zero calibration factor display Slope display Slope display Zero correction factor setting Zero correction factor setting Sensitivity correction factor setting Sensitivity correction factor setting Standard solution/check block selection Standard solution or check block Check block turbidity setting Check block turbidity Analog output range switching parameter setting Range switching output selection Range switching function selection Range A setting Zero point for range A Span point for range A Range B setting Zero point for range B Span point for range B Range C setting Zero point for range C Span point for range C Automatic range switching point Analog output 2 range selection Range selection for output 2 Fixed value for hold Fixed value for hold during failure CODE35 Hold during failure parameter setting Hold function enabled or disabled Last measured value, or fixed value Fixed value for output 1 Fixed value for output 2 IM 12E01A06-02E Display Setting Range Default * IN1 * IN2 * END * SPIKE * LIMIT * HLD-T * SMP-T 0: Disabled, 1: Enabled 000.000 to 999.999 005 to 600[s] 001 to 600[s] *CAL.A * SL *CAL.B –9.000 to 09.000[NTU] *CAL.K 0.2500 to 4.0000 *CAL.TP 0: Standard Solution, 1: Check Block 0:Disabled 999.999(NTU) 030(s) 030(s) Depends on MS Code 100.0(%) 00.000(NTU) 1.0000 1:Check Block *PLATE 0.001 to 2000[NTU] * RNGPR 0: Fixed, 1: AO1, 2: AO2 *RSET 0: Local, 1: Automatic, 2: Remote 2RNG, 3: Remote 3RNG * RSET.A * ZERO 0000.00 to 2000.00[NTU] *SPAN 0000.00 to 2000.00[NTU] * RSET.B *ZERO 0000.00 to 2000.00[NTU] * SPAN 0000.00 to 2000.00[NTU] * RSET.C * ZERO 0000.00 to 2000.00[NTU] * SPAN 0000.00 to 2000.00[NTU] *AUTOR 070 to 100[%] *mA2 0: 4.0-20.0mA, 1: 0.0-20.0mA 0:Fixed 1:Automatic 0000.00(NTU) 0010.00(NTU) 0000.00(NTU) 0100.00(NTU) 0000.00(NTU) 1000.00(NTU) 080(%) 0:4.0-20.0mA *H.mA2 00.0 to 22.0[mA](0-20mA) 02.0 to 22.0[mA](4-20mA) *FH.mA2 00.0 to 22.0[mA](0-20mA) 02.0 to 22.0[mA](4-20mA) 22.0(mA) * FHOLD 0: Disabled, 1: Enabled *FH.LST / *FH.LST(Last measured *FH.FIX value) *FH.FIX(Fixed value) *FH.mA1 02.0 to 22.0[mA] *FH.mA2 00.0 to 22.0[mA](0-20mA) 02.0 to 22.0[mA](4-20mA) 1: Enabled *FH.FIX:Fixed value 22.0(mA) 22.0(mA) 22.0(mA) 4-5 <4. PARAMETER SETTING> Table 4.3 Code CODE37 CODE40 CODE41 CODE44 CODE50 CODE51 CODE52 CODE54 CODE61 CODE64 CODE66 CODE67 CODE71 CODE72 Setting Parameters at Service Level (2/2) Setting Parameter Time constant setting Time constant in measurement Time constant in maintenance Contact output S1 function selection Function of contact output S1 Contact output S2 function selection Function of contact output S2 Delay time/hysteresis setting Delay time Hysteresis Auto return function selection Auto return function enabled or disabled Alarm setting function selection Alarm setting function enabled or disabled Password setting Password Negative nondisplay/nonoutput selection Negative nondisplay/nonoutput enabled or disabled Turbidity unit selection Turbidity unit Software version display Software version display Error “E2xx” error level selection Error levels of errors “E201” to “E206” Error “E3xx” error detection selection Error detection for “E301” to “E306” Error detection for “E307” Error detection for “E311” to “E316” Error detection for “E317” Analog output test Start test for output 1 Test current flowing End Start test for output 2 Test current flowing End Contact output test Test 1 Test 2 CODE73 CODE79 Contact input test Test Initialization Start Default loading Display Setting Range *TC 000 to 120[s] *TC-M 000 to 120[s] Default 020(s) 006(s) * S1 0:None , 1: Alarm, 2: Automatic calibration/ washing, 3: Maintenance 1: Alarm * S2 0:None , 1: Alarm, 2: Automatic calibration/ washing, 3: Maintenance 3: Maintenance * D.TIME 000 to 199[s] * HYST 000 to 100[%] 000(s) 002(%) *RET 0: Disabled, 1: Enabled 0: Disabled * MODE 0: Disabled, 1: Enabled 0: Disabled *PASS 0:No password, 1:111, 2:333, 3:777,4:888, 5:123, 6:957, 7:331, 8:546, 9:847 *MINUS 0: Disabled, 1: Enabled *UNIT 0: NTU, 1: FNU, 2: mg/l *VER 0.0.0 0.0: Disabled. Disabled 0: NTU 1.00 *201-6 0: Disable, 1: Level 1, 2: Level 2 1.1.2.2.2.2 *301-6 *307 *311-6 *317 1.1.1.1.1.1 1 1.1.1.1.1.1 1 0: Disabled, 1: Enabled 0: Disabled, 1: Enabled 0: Disabled, 1: Enabled 0: Disabled, 1: Enabled *AO1.T *END *AO2.T *END *DO.T1 0: Non-action, 1: Action *DO.T2 0: Non-action, 1: Action Last measured value held Last measured value held *DI.T *LOAD *WAIT IM 12E01A06-02E 4-6 <4. PARAMETER SETTING> 4.5 Procedures for Setting Parameters An example for setting parameters, the procedures are described in following modes. For more detail on the other functions, refer to User’s Manual IM 12E01A06-01E Section 6. (1) CAL Mode: {CALIB} (2) HOLD Mode: {HOLD} (3) RANGE Mode: {*RANGE} (1) CAL Mode: {CALIB} CALIB [YES] X.XXX X.XXX [NO] STD.CAL [YES] SMP.CAL [YES] *1 *2 X.XXX X.XXX [NO] ZERO [NO] SPAN [YES] *3 [NO] *4 *5 0.000 VALUE [ENT] X.XXX *5 [] [] [ENT] X.XXX ZERO [NO] ZRSPN [YES] 0.000 VALUE X.XXX VALUE.Z [] [] [ENT] *5 [YES] 0.000 VALUE ZERO Self-adjustment or [ENT] *6 X.XXX [NO] SPAN *5 [] [] [ENT] X.XXX SPAN [NO] ZERO [YES] 0.000 X.XXX X.XXX ZERO [YES] 0.000 VALUE [] [] [ENT] X.XXX SPAN X.XXX SPAN *5 [YES] *7 0.000 VALUE.S [] [] [ENT] Self-adjustment or [ENT] *6 [NO] X.XXX X.XXX SPAN Self-adjustment or [ENT] *6 Self-adjustment or [ENT] *6 Self-adjustment or [ENT] *6 Self-adjustment or [ENT] *6 CAL.END [YES] *7 HOLD Measurement mode *1: During the time between [YES] key press on {STD.CAL} display and [YES] or [NO] key press on {CAL.END} display or on error code display or [MODE] key press to abort, data display shows turbidity based on sensitivity correction factor (K)=1 and zero correction factor (B)=0. (Value not affected by SMP.CAL: Display by T1={(S0/(SL/100)}*(V-A), not by T2=K(T1+B)) *2: Span calibration in {STD.CAL} is performed using turbidity standards or check block that has been selected in CODE16. *3: In zero calibration, calibration value should be “0.000” and cannot be changed. Only [ENT] key is accepted to confirm. *4: In span calibration using check block, calibration value cannot be changed. Only [ENT] key is accepted to confirm. *5: First determine the decimal point using [>] key and then set value in the range of 0.000 to 20000 NTU. *6: Pressing [ENT] key during self-adjustment will abort the procedure, and the turbidity when [ENT] key is pressed will be used for calibration factor calculation. *7: Only [YES] key is accepted to confirm. Note: Calibration factor will be updated when the calibration procedure is completed without cancellation or errors. If [YES]/[NO] key is pressed on error code display, the instrument goes to {CALIB} display. Error E203 or E204 appears after the instrument once returns to measurement display. F4-4E.ai Figure 4.2 IM 12E01A06-02E CAL Mode Flow Chart <4. PARAMETER SETTING> (2) 4-7 HOLD Mode: {HOLD} In this mode whether the analog output is held or not in measurement mode, can be selected. This function is available when the hold function is set to “1: Enabled” in SET HOLD mode at the setting level. If the function is set to “0: Disabled,” this mode will be skipped. HOLD [YES] (Flashing) HOLD [NO] [YES] Hold disabled in measurement mode Hold enabled in measurement mode HOLD status indicator turns on. Measurement mode Figure 4.3 F4-5E.ai HOLD Mode Flow Chart If the hold function has been set to “1: Enabled” in SET HOLD mode at the setting level, the instrument automatically goes into this HOLD mode when it returns from any level to measurement mode, with the exception of DISPLAY mode. (3) RANGE Mode: {*RANGE} In this mode, an output range corresponding to analog output 1 and 2 can be set. Enter turbidity values corresponding to a 4 mA (or 0 mA) output signal (zero point) and a 20 mA output signal (span point). *RANGE [YES] *1 *FIXR.1 [NO] *2 *FIXR.2 [YES] XXXX.XX *ZERO [YES] [] [] XXXX.XX *ZERO [YES] XXXX.XX *SPAN *6 [ENT] [NO] [] [] [YES] [] [] XXXX.XX *SPAN *6 [] [] *3 *LOCAL [NO] [YES] *4 X [] Local range for output 1 X: 0(Range A), 1(Range B) or 2(Range C) [] Local range for output 2 X: 0(Range A), 1(Range B) or 2(Range C) *LCL.1 [YES] *5 X *LCL.2 [ENT] [ENT] *1: Skipped if analog output range switching selection {*RNGPR} is set to “1: Analog output 1” in CODE 30 at service level. *2: Skipped if analog output range switching selection {*RNGPR} is set to “2: Analog output 2” in CODE 30 at service level. *3: Displayed only if analog output range switching selection {*RNGPR} is set to “1: Analog output 1” or “2: Analog output 2” and range switching function selection {*RSET} is set to “0: Manual” in CODE 30 at service level. *4: Displayed only if analog output range switching selection {*RNGPR} is set to “1: Analog output 1” in CODE 30 at service level. *5: Displayed only if analog output range switching selection {*RNGPR} is set to “2: Analog output 2” in CODE 30 at service level. *6: Zero and span points should be set together. Only after span point has been entered, both zero and span points are accepted. Cancellation by pressing [MODE] key during the procedure will not change either of the points. F4-6E.ai Figure 4.4 RANGE Mode Flow Chart IM 12E01A06-02E Blank Page 5-1 <5. MAINTENANCE> 5. MAINTENANCE 5.1 Lamp Replacement CAUTION Wait for some time to replace lamp assembly because lamp temperature is high just after lamp power was turned off. Although the lamp has a normal service life of more than one year, it is recommended that the lamp be replaced at an interval of one year in terms of preventive maintenance. Some lamps may fail within one year. The following describes how to replace the lamp. (1) Remove power from the TB750G turbidimeter. (2) Remove the cover of the light source unit on the left side of the detector by loosening the 4 setscrews. (3) Disconnect the terminals for lamp from the terminal block. Remove the lamp assembly by loosening the 2 setscrews. (4) Replace the lamp assembly with a new one. Mount the new lamp assembly with the 2 setscrews. Connect the terminals for lamp to the terminal block (LAMP). (5) Supply power to the TB750G turbidimeter. (6) Check that the lamp is lit. Replace and fix the cover of the light source unit by tightening the 4 setscrews. (7) After the lamp replacement, allow the turbidimeter to warm up for at least one hour. Then, perform calibrations instructed in Sections 3. (8) Lamp replacement is now complete. Cover of the Light Source Unit Lamp Assembly A Cover for Setscrews (2) Terminal Terminal Block Setscrews (2) Terminal Block Lamp Assembly [View A] Figure 5.1 F5-1E.ai Lamp Replacement IM 12E01A06-02E 5-2 <5. MAINTENANCE> 5.2 Calibration 5.2.1 Outline (1) CALIB STD.CAL SMP.CAL ZERO Zero calibration using zero turbidity water SPAN Span calibration using check block Span calibration using standard solution ZRSPN 2-point calibration using standard solutions ZERO Zero point correction SPAN Sensitivity correction (2) (3) Figure 5.2 Calibration using filtered water as zero reference F5-2E.ai Calibration Methods (1) Select when performing a calibration using 0.2 or 1.0 µm filtered water as zero reference. (2) Select when performing a 2-point calibration using standard solutions. Use this method to perform a calibration complying with EPA 180.1. (3) Select when performing a zero point or sensitivity correction after calibration (1) or (2). This method is used to adjust the meter reading to reflect the lab data in a grab sample calibration. Note: The instrument has been pre-calibrated using 0.2 µm filtered water as zero reference at the factory before shipment. 5.2.2 2-point Calibration Using Turbidity Standard Solutions A 2-point calibration of the TB750G is performed with turbidity standard solutions equivalent to zero solution and to span solution. After accepting both zero and span values, the meter calculates a calibration factor. When user-prepared zero and span solutions are used for calibrations, the turbidity of diluting water for calibration should be measured with a calibrated lab turbidimeter in advance. Prepare a formazin standard solution as a span solution. If the turbidity of diluting water is high against the measurement accuracy, add the turbidity of diluting water to the nominal turbidity of the prepared formazin standard solution. Use diluting water as zero solution and a formazin solution as span solution for the calibration. Refer to Section 3, “Operation” for concrete operating procedures. IM 12E01A06-02E 6-1 <6. TROUBLESHOOTING> 6. TROUBLESHOOTING Table 6.1 Error Code List (1/3) No. Error Occurrence Problem E101 Flash memory failure When power is turned on. SUM results of flash memory do not match programmed SUM values 5 consecutive times. E102 EEPROM write failure All modes Data is written in EEPROM and verified, and write failure occurs 3 consecutive times. E103 RAM failure When power is turned on. RAM area failure (3 consecutive times). E104 AD converter failure All modes AD converter failure (3 consecutive times). E201 Input voltage failure All modes Either IN1 or IN2 input voltage is outside the range of -0.15 to 1.2 V (fixed), including abnormal value 0x7FFF or 8000, for 5 consecutive seconds. E202 Detection element failure All modes Either IN1 or IN2 input voltage is less than specified voltage for 5 consecutive seconds. E203 Lamp life expired E204 Lamp intensity failure Zero calibration coefficient A rewriting timing. (excluding when E301 or E311 occurs.)*1 During zero calibration self-adjustment, input voltage IN1 is less than specified voltage for 5 consecutive seconds. (Not detected when E301 is occurring.) Rewriting of zero calibration coefficient A is performed. E205 Calibration failure E301 to E307, E311 to E317, E321 during PC communication occurrence timing Any of E301 to E307, E321 during PC communication is occurring. Return When power is turned off and then on again, and problem is eliminated. (Without [YES]/[NO] key press.) When problem is eliminated for 5 consecutive seconds. When [YES]/[NO] key is pressed, error code display disappears, at the same time during failure, error is once removed. When problem is eliminated. When [YES]/[NO] key is pressed, error code display disappears, at the same time during failure, error is once removed. IM 12E01A06-02E 6-2 <6. TROUBLESHOOTING> Table 6.1 No. Error Code List (2/3) Error Occurrence Problem Return When problem is eliminated. When [YES]/[NO] key is pressed, error code display disappears, at the same time during failure, error is once removed. E301 Turbidity zero calibration coefficient A failure During zero calibration (zero water). Turbidity zero calibration coefficient A newly determined after zero calibration (zero water or lamp OFF) is outside of the range of 0 to 5, or each coordinate value of newly determined linearized table is outside of the range of -99999 to 999999. E302 Turbidity slope SL failure During span calibration (standard solution) Turbidity slope value SL newly determined after span calibration (standard solution) is outside of the range of 25 to 200%, or each coordinate value of newly determined linearized table is outside of the range of -99999 to 999999. E303 Turbidity check block failure During span calibration (check block) Turbidity slope value SL newly determined after span calibration (check block) is outside of the range of 50 to 150%, or each coordinate value of newly determined linearized table is outside of the range of -99999 to 999999. E304 Turbidity zero correction factor B failure During zero shift calibration Turbidity zero correction factor B newly determined after zero shift calibration is outside of the range of -9 to 9. E305 Turbidity sensitivity correction factor K failure During span calibration (sensitivity correction) Turbidity sensitivity correction factor K newly determined after span calibration (sensitivity correction) is outside of the range of 0.25 to 4. E306 Turbidity reference sensitivity SO failure During turbidity reference sensitivity calibration Turbidity reference sensitivity SO newly set after span calibration (check block) is outside of the tolerance of 0.0001 to 2000, or each coordinate value of newly determined linearized table is outside of the range of -99999 to 999999. E307 Turbidity response time failure All calibrations Turbidity self-adjustment is not complete before self- adjustment time has elapsed. E321 Communication During error communication with PC IM 12E01A06-02E Any communication error during communication with PC. When problem is eliminated in the next communication (only one time is OK). When [YES]/[NO] key is pressed, error code display disappears, at the same time during failure, error is once removed. (excluding RS signal during communication with calibration/cleaning controller.) 6-3 <6. TROUBLESHOOTING> Table 6.1 No. Error Code List (3/3) Error Occurrence Problem Return When [YES]/[NO] key is pressed, error code display disappears and error is removed. E351 Analog output range setting failure During parameter setting in RANGE mode or CODE 30. Either of: (1) Zero point ≥ span point, or (2) Span point - zero point < 20% of span point or 0.20, whichever is greater (for turbidity) or ( span point - zero point < 5.00 (for color). When automatic range is selected in CODE 30, also possible: (3) At each span point, Range A ≥ Range B, or (4) at each span point, Range B ≥ Range C E352 Parameter setting failure During parameter setting Value outside the setting range is set in setting other than AO range setting. Refer to the user’s manual IM 12E01A06-01E for details. IM 12E01A06-02E Blank Page i Revision Information Model TB750G Right Angle Scattered Light Turbidimeter Quick Start Manual  Title:  Manual No.: IM 12E01A06-02E Edition Date Remark (s) 2nd Mar. 2015 P.2-3, P.6-1, Some revision of Section 2.2.1, “Wiring Required for TB750G” : Note 6 is added to figure 2.3. Some revision of Section 6, “TROUBLESHOOTING” : The name of error code E202 is revised, and “PD check voltage” and “lamp life check voltage” changed to “specified voltage”. 1st Newly published Apr. 2006 IM 12E01A06-02E Blank Page