Yokogawa Model TB750G Right Angle Scattered Light Turbidimeter Quick Start Manual
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
User’s
Manual
Model TB750G
Right Angle Scattered Light
Turbidimeter
Quick Start Manual
IM 12E01A06-02E
R
IM 12E01A06-02E
2nd Edition
i
INTRODUCTION
n Structure of this Manual
This is a supplementary manual to TB750G Right Angle Scattered Light Turbidimeter for User's Manual
IM 12E01A06-01E. This manual describes the specifications, installation, operation, maintenance, and
troubleshooting. To use this instrument correctly, read this manual and User's Manual IM 12E01A0601E thoroughly.
n Before Measurement
The TB750G turbidimeter is preset with defaults prior factory shipment. Before measurement, verify that
these factory default settings meet the operating conditions and if necessary, reconfigure parameters.
Symbol and Notations Used in this Manual
n Symbol
The following symbol is used in this manual.
CAUTION
Indicates that damage to software or hardware, or system failures could occur if instructions are not
followed.
n Notations
In this manual, operation keys, display messages, indicators, and LED lamps on the instrument are
denoted in the following manner.
1. Operation keys are denoted by brackets [ ] – for example:
YES
: [YES] key
2. Indicators, messages, and numerical data on the display are denoted as follows:
Message display: {*WASH}
Data display: {2.05} (when lit), {2.05} (when flashing)
Media No.IM 12E01A06-02E (CD)
2nd Edition :Mar. 2015 (YK)
All Rights Reserved Copyright © 2006, Yokogawa Electric Corporation
IM 12E01A06-02E
ii
Notice
n About this Manual
l This manual should be passed on to the end user.
l This manual should be read thoroughly before operating the instrument.
l This manual explains the functions contained in this product, but does not warrant that they will suit
the particular purpose of the user.
l The contents of this manual shall not be reproduced or copied, in part or in whole, without permission.
l The contents of this manual are subject to change without prior notice.
l Every effort has been made to ensure the accuracy in the preparation of this manual. However, if
any errors or omissions are noticed, please contact the nearest Yokogawa representative or sales
office.
n Protection, Safety, and Modification of the Product
l The safety instructions described in this manual should be strictly observed to ensure safety both
of the product and the system controlled by the product.
l A protection or safety circuit should be installed externally, if needed. Do not attempt to modify or
add such circuit to the inside of the equipment.
n Limitation of Liability
l Yokogawa grants no warranties other than the express warranty set forth under the warranty provisions.
l Yokogawa shall not be liable to you or any third party for any damage, including consequential or
incidental damages, arising out of or in connection with the use of this equipment, defects beyond
our knowledge, or any other contingency beyond our control.
IM 12E01A06-02E
Toc-1
Model TB750G
Right Angle Scattered Light Turbidimeter Quick Start Manual
IM 12E01A06-02E 1st Edition
CONTENTS
INTRODUCTION........................................................................................................i
Symbol and Notations Used in this Manual.................................................................. i
Notice................................................................................................................................ ii
1. OVERVIEW................................................................................................. 1-1
1.1 Features.............................................................................................................. 1-1
1.2
2.
Measurement Principle..................................................................................... 1-1
PIPING AND WIRING................................................................................. 2-1
2.1 Piping.................................................................................................................. 2-1
2.1.1
System Using Open Head Tank and Zero Turbidity Filter.................. 2-1
2.1.2
System without Head Tank and Zero Turbidity Filter.......................... 2-2
2.2 Wiring.................................................................................................................. 2-3
2.2.1
Wiring Required for TB750G.............................................................. 2-3
2.2.2
Wiring for Detector.............................................................................. 2-4
2.2.3
Wiring for Converter............................................................................ 2-5
3. OPERATION............................................................................................... 3-1
3.1 Preparation......................................................................................................... 3-1
3.1.1 Outline................................................................................................. 3-1
3.2
3.1.2
Converter Operation Panel................................................................. 3-1
3.1.3
Calibration........................................................................................... 3-2
3.1.4
Performing Zero/span Calibration Using Filtered Water as
Zero Reference................................................................................... 3-3
3.1.5
Performing 2-point Calibration Using Standard Solution.................... 3-7
Normal Operation............................................................................................... 3-8
3.2.1
When Sample Water Supply is Cut Off............................................... 3-8
3.2.2 Shutdown............................................................................................ 3-8
3.2.3 Restart................................................................................................. 3-8
4.
PARAMETER SETTING............................................................................. 4-1
4.1
Outline of Operation, Setting, and Service Levels......................................... 4-1
4.2
Operation Level.................................................................................................. 4-2
4.3
Setting Level....................................................................................................... 4-3
4.4
Service Level...................................................................................................... 4-4
4.5
Procedures for Setting Parameters................................................................. 4-6
IM 12E01A06-02E
Toc-2
5. MAINTENANCE.......................................................................................... 5-1
5.1
Lamp Replacement............................................................................................ 5-1
5.2 Calibration.......................................................................................................... 5-2
5.2.1 Outline................................................................................................. 5-2
5.2.2
2-point Calibration Using Turbidity Standard Solutions...................... 5-2
6. TROUBLESHOOTING............................................................................... 6-1
Revision Information................................................................................................i
IM 12E01A06-02E
1-1
<1.Overview>
1. OVERVIEW
1.1 Features
The TB750G is a process turbidimeter employing the right angle light scattering method, and has the
following features.
• Highly reliable measurement with excellent linearity and repeatability
- Linearity: ±2% of reading or ±0.01 NTU, whichever is greater
- Repeatability: ±1% of reading or ±0.002 NTU, whichever is greater
- Display resolution: 0.001 NTU
• Easy-to-clean measurement cell
• Compact, lightweight converter and detector
• User configurable analog output range
- Analog output range: 0-0.2 NTU to 0-100 NTU
• Analog output range switching (2 or 3 ranges)
• Enhanced self-diagnostic function as standard
- Light source failure, input element failure, calibration failure, various circuit failures, etc.
• Detector designed to remove influence of air bubbles
• A wide range of measurement conditions
- Low flow rate: 0.05 to 20 l/min {0.8 to 317 gal/h}
- High pressure: Up to 500 kPa {72 psi}
- Sample temperature: 0 to 50°C {32 to 122°F}
• Detector can be connected for in-line analysis
• 2 analog outputs, 3 relay contact outputs, and 1 serial communication
• Many options available upon request
• Ultrasonic transducer and oscillator for ultrasonic cleaning
• Various head tanks to accommodate application requirements
1.2
Measurement Principle
The measuring system of the TB750G turbidimeter, comprised of a detector and a converter, complys
with USEPA 180.1.
Scatted Light Detection Element
Scatted Light Detector
Receiver (Darkroom)
Light Source Lens
(Lamp)
Measuring
Water
Measurement Window (Glass)
F1-1E.ai
Figure 1.1
Measurement Principle
IM 12E01A06-02E
Blank Page
2.
2-1
<2. PIPING AND WIRING>
PIPING AND WIRING
2.1 Piping
Piping methods depend on the application, select the appropriate system configuration, and install
pipes accordingly following examples.
2.1.1 System Using Open Head Tank and Zero Turbidity Filter
This is a typical system. A piping diagram is shown in Figure 2.1. Air bubbles in a water sample are
removed by an open head tank and the water sample is introduced into the detector at a stable flow
rate.
Head Tank
H
Flow Control Valve
Sample Water
Outlet
Sample Water Drain
Valve
Valve
Drain
Tap
Water
Drain
TB750G
Detector
h
Sample
Water
Sample Water
Effluent
Drain
Sample Water Sample Water Drain Port
Inlet
Sample Water
Drain Valve
Supply Valve
Zero Water
Drain
Zero Water Supply Valve
Tap Water
Valve
Zero Water
Drain Valve
Zero Turbidity
Filter
Other Turbidity Detectors
F2-1E.ai
Figure 2.1
Piping Diagram
• To prevent corrosion of water inside the zero turbidity filter, allow water to flow through the filter
and flow out from the zero water drain valve continuously at a flow rate of approximately 10 ml/
min.
• When using filtered water as zero reference, install a filter with the appropriate pore size
depending on the measuring range.
Measuring range of less than 2 NTU:
0.2 µm
Measuring range of 2 NTU or greater:
1 µm
IM 12E01A06-02E
2-2
<2. PIPING AND WIRING>
2.1.2 System without Head Tank and Zero Turbidity Filter
This is a simple system where a water sample is taken from the process and directly introduced
into the detector. This system configuration can be used when a water sample contains a negligible
amount of air bubbles or when the turbidity of a water sample is high and the effect of air bubbles is
nonsignificant.
CAUTION
Sample water conditions and ambient temperature must meet the specifications: maximum pressure:
500 kPa; sample temperature: 0 to 50°C; ambient temperature: -5 to 50°C.
Flow Control Valve
Sample Water
Effluent
Sample Water
Outlet
Sample
Water
Sample Water
Sample Water
Valve
Tap Water
Zero Water
Sample Water
Inlet
Sample Water
Supply Valve
h
TB750G
Detector
Process Piping
Drain
Drain Port
Drain Valve
Drain
Zero Water Supply Valve
Tap Water
Valve
Zero Water
Drain Valve
Zero Turbidity
Filter
Other Turbidity Detectors
F2-2E.ai
Figure 2.2
Piping Diagram
• The flow control should not be done on the inlet side of the detector.
• To prevent corrosion inside the zero turbidity filter, allow water to flow through the filter and flow
out from the zero water drain valve continuously at a flow rate of approximately 10 ml/min.
IM 12E01A06-02E
2-3
<2. PIPING AND WIRING>
2.2 Wiring
2.2.1 Wiring Required for TB750G
TB750G Turbidity detector
TB750G Turbidity converter
L
CONVERTN POWER
ER
G
G
Power
supply *5
Grounding *1
(100Ω or less)
L
N TUS
G
Dedicated cable (1/2/3m)
U1U2
Grounding *1
(100Ω or less)
Power supply
cable *6
TUS400G Ultrasonic Oscillator *5
L2
L1
SENSOR
S1
NO
NC
COM
Contact
output S1
S2
NO
NC
COM
Contact
output S2
FAIL
NO
NC
COM
Contact
output FAIL
RANGE
A
B
C
COM
Range
contact output
G
Serial communication *3 *4
(RS-422)
Ultrasonic transducer
connecting cable
(maximum length: 15m)
(customer wiring)
Contact input
U1
U2
U3
S
Analog output 1
(4-20mA DC) *3
Analog output 2
(4-20mA DC or
0-20mA DC) *3
Grounding *2
(100Ω or less)
RX+
RXTX+ RS-422
TXSG
G
IN1
IN2
COM
+
+
G
INPUT
mA1
mA2
OUTPUT
Grounding *1
(100Ω or less)
(Note) Dotted wiring is external wiring. Use cable with 6 to 12 mm OD for wiring.
*1 Power terminal "G" on detector, detector case, and converter case must be grounded (ground resistance: 100Ω or less).
*2 External grounding terminal of ultrasonic oscillator must be grounded (ground resistance: 100Ω or less).
*3 Use 2-conductor shielded cable for analog output wiring and serial communication wiring.
*4 The wiring configuration is described below in case that RS-232C serial communication is selected.
Serial communication
(RS-232C)
*5
*6
TXD
RXD RS-232C
SG
G
When option code "/US" is specified, TUS400G should be purchased separately.
When TUS400G is used in system, the power supply to TB750G should be the same as the supply voltage specified in the
MS Code of TUS400G.
When "TB750G/US" and "TUS400G-NN-RC" or "TUS400G-NN-KC" is specified, refer to IM 19C1B4-01E for wiring.
Figure 2.3
F2-3E.ai
Connection Diagram
IM 12E01A06-02E
2-4
<2. PIPING AND WIRING>
2.2.2 Wiring for Detector
Wire for connections to the following terminals:
(1) Power supply terminals;
(2) Power supply terminals for ultrasonic oscillator (when option code “/US” is specified); and,
(3) External grounding terminal.
Terminals TM3 (Power to Lamp)
(1) Power Supply Terminals
(2) Power Supply Terminals for
Ultrasonic Oscillator
Figure 2.4
Terminal G
(For Converter - Detector
Connecting Cable)
Connector for Converter - Detector Cable
F2-4E.ai
Internal View of Detector
(3) External Grounding Terminal (M5 Screw)
(Norminal size of at least 2 mm2)
F2-5E.ai
Figure 2.5
IM 12E01A06-02E
Grounding Terminal of Detector
2-5
<2. PIPING AND WIRING>
2.2.3 Wiring for Converter
Wire for connections to the following terminals:
(1) Analog output/contact input terminals;
(2) Contact output/range contact output terminals;
(3) Serial communication terminals (when in use); and,
(4) Grounding terminal.
(A) Converter - Detector
Connecting Cable
Front Cover
Serial Communication
Terminals
Connector for
Converter - Detector Cable
(3) Serial Communication
Front Cover
Terminals
Terminal Cover
(B)
Terminal G
(For Converter Detector Connecting Cable)
(1) Analog Output/Contact Input Terminals
(2) Contact Output/
Range Contact
Output Terminals
Analog Output/Contact Input Terminals
F2-6E.ai
How to Remove the Terminal Cover
Disconnect the converter-detector connecting cable (A) and its grounding wire (B) from the converter. Remove the terminal cover by
sliding it slightly to the left and pulling it forward.
Figure 2.6
Internal View of Converter
Toothed Lock Washer
(4) Grounding Terminal (M4 Screw)
Grounding Wire
(Nominal Size of at least 2 mm2)
F2-7E.ai
Figure 2.7
Grounding Terminal of Converter
IM 12E01A06-02E
2-6
<2. PIPING AND WIRING>
Contact Output (S1, S2 and FAIL) Wiring
• Be sure to use the contacts meeting the conditions below.
• Functions of contact outputs S1 and S2 should be set in Codes 40 and 41 at service level. Refer
to section 4.3.
Table 2.1
Contact Rating
AC
DC
Contact maximum permissible voltage
250 V
30 V
Contact maximum permissible current
2A
3A
Contact maximum permissible power
(resistance load)
125 VA
60 W
Table 2.2
Contact Operation
Contact S1, S2
Status
LED
In action
ON
Not in action
OFF
Open Closed OFF Closed Open
Power OFF
OFF
Open Closed OFF
IM 12E01A06-02E
NO
NC
Contact FAIL
Closed Open
LED
ON
NO
NC
Open Closed
Open Closed
<3. OPERATION>
3-1
3. OPERATION
3.1 Preparation
3.1.1 Outline
• Make sure that TB750G converter and detector
are fixed fimly.
• Unused cable inlets should be plugged to
prevent moisture from entering.
1. Installation Check
2. Piping Check
3. Wiring Check
• Note that the TB750G does not have an internal
power switch.
4. Supplying Power
• Upon power up, allow the meter to warm up for
at least one hour.
• When using zero turbidity filter, allow water to
run through it for at least one hour.
5. Warm-up
6. Setting and Checking Parameters
• For details of parameter settings, refer to
chapter 4.
7. Calibration
• For details of calibrations, refer to the following
pages.
8. Supplying a Water Sample and
adjusting the flow rate
9. Operation check
• Adjust the opening of the valve at the sample
water out of the detector.
• Make sure that the system is free of defects.
F3-1E.ai
3.1.2 Converter Operation Panel
Turbidity Unit Indication
Operation Key
([*]: To enter the operation level
from measurement mode)
Contact Status LED
LCD Screen
Operation Key
F3-2E.ai
IM 12E01A06-02E
3-2
<3. OPERATION>
3.1.3 Calibration
(1) When performing a calibration using 0.2 or 1.0 µm filtered water as zero reference, refer to
section 3.1.4.
(2) When performing a 2-point calibration using standard solutions, refer to section 3.1.5. Use this
method to perform a calibration complying with EPA 180.1.
Note: The instrument has been pre-calibrated using 0.2 µm filtered water as zero reference at the
factory before shipment.
Calibration Flow Chart
Start calibration
NO
Is filtered water used
as zero reference?
YES
Perform span/zero calibration
using filtered water
Perform 2-point calibration
using standard solutions*
Perform zero calibration
Perform point calibration
first using zero solution
Is span calibration performed
using check block or
standard solution?
Standard Solution
Check block
Perform span calibration
using check block.
Refer to Section 3.1.4.
Perform span calibration
using standard solution.
Refer to Section 3.1.4.
Perform second point
calibration using span solution.
Refer to Section 3.1.5.
End calibration
* Calibration complying with EPA Method 180.1.
IM 12E01A06-02E
F3-3E.ai
3-3
<3. OPERATION>
3.1.4 Performing Zero/span Calibration Using Filtered Water as Zero Reference
START
• Supply power
• Warm-up (at least 1 hour)
• Allow water to run through the zero filter
• Supply zero water
• Prepare span solution, if necessary
[MODE]
CALB
[YES]
X.XXX
STD.CAL
Automatically goes into
maintenance mode and holds
the output.
[YES]
X.XXX
ZERO
• Confirm reading has stabilized (3 to 5 min)
[YES]
0.000
VALUE
{0.000} None configurable
[ENT]
X.XXX
ZERO
X.XXX
CAL.END
Auto-adjustment
Zero calibration is complete
[NO]
A
To span calibration
using check block
[YES]
B
To span calibration
using standard solution
F3-4E.ai
IM 12E01A06-02E
3-4
<3. OPERATION>
Span calibration using check block
Note: For span calibration using standard solution, refer to next page.
From zero calibration
A
X.XXX
STD.CAL
[YES]
X.XXX
ZERO
[NO]
X.XXX
SPAN
• Stop supplying zero water
• Drain zero water
• Install check block *
• Supply zero water
• Confirm reading has stabilized (3-5 min)
[YES]
XX.XX
VALUE
{XX.XX} None configurable
• Check that display reads the value of check block
[ENT]
XX.XX
SPAN
XX.XX
CAL.END
Auto-adjustment
Span calibration is complete
[YES]
HOLD
Release HOLD **
[NO]
END
* How to install check block
1. Remove the rubber cover from the top of the detector.
2. Remove the retaining plate on the window of the top cover
by loosening the 2 setscrews.
3. Detach the observation window and replace it with the
check block. Fix it with the 2 setscrews securely.
Check block
** Be sure to release HOLD
IM 12E01A06-02E
F3-5E.ai
<3. OPERATION>
3-5
Span calibration using standard solution (1/2)
From zero calibration
Before performing span calibration using
standard solution, standard solution/check
block selection should be set to "0 (standard
solution)" in CODE 16 as shown below.
B
HOLD
[YES]
HOLD
[*] at least second
To enter operation level
*SETP
[NO]
[NO]
[NO]
(3 times)
{*SETP} -> {*RANGE} -> {*HOLD} -> {*SERV}
*SERV
[YES]
16
*CODE
[]
[]
• Enter "16" using []and [] keys
[YES]
0
*CAL.TP
[]
• Enter "0" using [] key
[ENT]
*SERV
[MODE]
HOLD
[YES]
HOLD
[MODE]
C
F3-6E.ai
IM 12E01A06-02E
3-6
<3. OPERATION>
Span calibration using standard solution (2/2)
C
CALB
[YES]
X.XXX
STD.CAL
[YES]
X.XXX
ZERO
[NO]
X.XXX
SPAN
• Remove top cover (6 setscrews)
• Rinse the cell with span solution and drain
• Fill the cell with span solution to 90% *
• Replace top cover with setscrews
• Confirm reading has stabilized (3-5 min)
[YES]
00.00
VALUE
[]
• Set decimal point using [] key
[ENT]
20.00 **
VALUE
[]
[]
• Enter value of span calibration using [] and [] keys
[ENT]
XX.XX
SPAN
XX.XX
CAL.END
Auto-adjustment
Span calibration is complete
[YES]
HOLD
Release HOLD ***
[NO]
END
* How to pour span solution
Pour span solution into the measurement cell very
slowly, using a funnel or relevant equipment, taking
care not to allow air bubbles to enter the cell.
** Number in the figure is an example.
*** Be sure to release HOLD
IM 12E01A06-02E
F3-7E.ai
<3. OPERATION>
3-7
3.1.5 Performing 2-point Calibration Using Standard Solution
START
• Supply power
• Warm-up (at least 1 hour)
• Wash the cell
• Prepare zero and span solutions
[MODE]
CALB
[YES]
X.XXX
STD.CAL
[NO]
X.XXX
SMP.CAL
[YES]
X.XXX
ZRSPAN
•
•
•
•
•
Remove top cover (6 setscrews)
Rinse the cell with zero solution and drain
Fill the cell with zero solution to 90% *
Replace top cover with screws
Confirm reading has stabilized (3-5 min)
[YES]
0.000
VALUE.Z
[]
• Set decimal point using [] key
[ENT]
0.020 **
VALUE.Z
[]
[]
• Enter value of zero calibration using [] and [] keys
[ENT]
X.XXX
ZERO
Auto-adjustment
X.XXX
SPAN
• Remove top cover (6 setscrews)
• Rinse the cell with zero solution and drain
• Fill the cell with zero solution to 90% *
• Replace top cover with screws
• Confirm reading has stabilized (3-5 min)
[YES]
00.00
VALUE.S
[]
• Set decimal point using [] key
[ENT]
20.00 **
VALUE.S
[]
[]
• Enter value of span calibration using [] and [] keys
[ENT]
XX.XX
SPAN
XX.XX
CAL.END
Auto-adjustment
Span calibration is complete
[YES]
HOLD
Release HOLD ***
[NO]
END
* How to pour zero solution
Pour a span solution into the measurement cell very
slowly, using a funnel or relevant equipment, taking
care not to allow air bubbles to enter the cell.
** Numbers in the figures are examples.
*** Be sure to release HOLD
F3-8E.ai
IM 12E01A06-02E
3-8
<3. OPERATION>
3.2
Normal Operation
The TB750G turbidimeter does not normally require any manual operations except when periodic
cleaning/calibration is performed or when a failure occurs.
3.2.1 When Sample Water Supply is Cut Off
The turbidity measurement is not affected by temporary suspension of sample water supply, but longterm suspension may interfere with the accuracy of the instrument. Since the TB750G cannot detect
whether the sample water supply is stopped or not, periodical check of the sample line is needed.
3.2.2 Shutdown
The parameter settings and other information set in the turbidimeter are retained even if power is
turned off. If the instrument will be out of operation for a long period, power should be removed.
Dirt and/or stains on the instrument should be washed off thoroughly when the turbidimeter is to be
removed from the site. Rinse the inside of the measurement cell, and then empty the cell or keep zero
water running through the cell.
3.2.3 Restart
When power is returned, the turbidimeter is put into the measurement state. It requires at least one
hour for warm-up. Check that the reading stabilizes well, before performing calibration.
IM 12E01A06-02E
4-1
<4. PARAMETER SETTING>
4.
PARAMETER SETTING
4.1
Outline of Operation, Setting, and Service Levels
Parameters should be set in the relevant mode classified into three levels: operation, setting and
service.
Measurement Mode
[MODE] key
Operation Level
[ * ] For at least second
Setting Level
CALIB (Calibration)
STEP (Alarm setting)
DISP (Message display)
RANGE (Analog output)
STEP (Alarm setting)
HOLD (Hold selection)
HOLD (Hold selection)
Service Level
CODE XX
Figure 4.1
F4-1E.ai
Transition between Measurement Mode and Three Levels
IM 12E01A06-02E
4-2
<4. PARAMETER SETTING>
4.2
Operation Level
This is the level where, basically, operations regarding routine inspections and maintenance, such as
calibration and manual cleaning, can be done. Also, the desired display item on the message display
can be selected at this operation level.
Measurement
mode
[MODE]
CALIB
[YES]
[YES]
Calibration mode
HOLD
[NO]
DISP
[YES]
[YES]
Message display selection mode
[NO]
*1
SETP
[YES]
[YES]
Alarm setting mode
[NO]
*2
HOLD
[YES]
HOLD
*3
[YES]
Hold selection mode
[NO]
*1: Skipped if alarm setting function is set to “0: Disabled” in Code 51 at service level, or
unless either of functions of contact output S1 or S2 is set to “1: Alarm” in Code 40 or 41
at service level.
*2: Skipped if hold function is set to “*H.OFF: Disabled” in SET HOLD mode at setting level.
*3: If [YES] key is pressed, analog output is held in measurement mode.
F4-2E.ai
Table 4.1
Setting Parameters at Operation Level
Mode
CAL
Setting Parameter
Calibration
Using turbidity standards,
or using water sample
Start
Turbidity value
Self-adjustment in process
End
Display
Default
STD.CAL or
SMP.CAL
ZERO or SPAN
or ZRSPN
VALUE
0.001 to 2000[NTU]
*1
ZERO or SPAN
CAL.END
CAL START
Not available
WASH START
DISPLAY
Setting Range
CALIB
Not available
Message display selection
Analog output 1 current
Analog output 2 current
Switchable output range
state
DISP
XX.XmA1
Output 1 current
XX.XmA2
RNG1/2-A/B/C
*2
* SETP
*3
(No pointer at
Alarm setting
mode display)
High alarm setpoint
* S.TB-H
–010.0 to 2200.0[NTU] *1
2200.0(NTU)
Low alarm setpoint
* S.TB-L
–010.0 to 2200.0[NTU] *1
-010.0(NTU)
HOLD
Hold selection
Hold selection
*1:
*2:
*3:
*4:
IM 12E01A06-02E
HOLD
*4
(Disabled)
Although setting range is up to 2000 NTU, measuring range is up to 100 NTU.
Skipped if analog output range switching selection {*RNGPR} is set to “0: Fixed” in CODE 30 at service level.
Skipped if alarm setting function is set to “0: Disabled” in Code 51 at service level, or if either function of contact output S1 or S2
is not set to “1: Alarm” in Code 40 or 41 at service level.
Skipped if hold function is set to “*H.OFF: Disabled” in SET HOLD mode at setting level.
4-3
<4. PARAMETER SETTING>
4.3
Setting Level
There are modes where parameters regarding output signals and contact outputs can be set.
*2
Measurement
mode
HOLD
[ * ] For at least second
*1
*SETP
[YES]
SETPOINTS Mode
[NO]
*RANGE
[YES]
RANGE Mode
[NO]
*HOLD
[NO]
*SERV
[YES]
*3
[YES]
SET HOLD Mode
Service Level
[NO]
*1: Skipped unless either of functions of contact output S1 or S2 is set to “1:
Alarm” in Code 40 or 41 at service level.
*2: Skipped if hold function is set to “*H.OFF: Disabled” in SET HOLD mode at
setting level.
*3: For service level, see Section 4.4.
Table 4.2
Mode
Setting Parameters at Setting Level
Setting Parameter
SETPOINTS Alarm setting
RANGE
Display
* SETP
Setting Range
*1
High alarm setpoint
* S.TB-H
–010.0 to 2200.0[NTU]
*1 2200.0(NTU)
* S.TB-L
–010.0 to 2200.0[NTU]
*1 -010.0(NTU)
Analog output range
setting
* RANGE
Zero point
Span point
Local range selection
Local range for output 1,
or local range for output 2
Hold parameter setting
* FIXR.1 or
*FIXR.2
* ZERO
0000.00 to 2000.00[NTU]
0000.00(NTU)
* SPAN
0000.00 to 2000.00[NTU]
0100.00(NTU)
* LOCAL
*LCL.1 or
*LCL.2
0: Range A, 1: Range B, 2: Range C 0:Range A
* HOLD
Hold function enabled, or
hold function disabled
* H.ON or
*H.OFF
Last measured value, or
fixed value
* H.LST or
* H.FIX
*H.ON (Enabled)
*H.OFF (Disabled)
* H.ON:Enabled
*H.LST (Last measured value)
*H.FIX (Fixed value)
* H.LST:Last
measured value
Fixed value for output 1
* H.mA1
02.0 to 22.0[mA]
22.0(mA)
Fixed value for output 2
* H.mA2
00.0 to 22.0[mA](0-20mA)
02.0 to 22.0[mA](4-20mA)
22.0(mA)
CAL/WASH
Not available
SERVICE
*1:
*2:
Default
Low alarm setpoint
Fixed range for output 1,
or fixed range for output 2
SET HOLD
F4-3E.ai
*2
Skipped if alarm setting function is set to “0: Disabled” in Code 51 at service level, or if either function of contact output S1 or S2
is not set to “1: alarm” in Code 40 or 41 at service level.
For service level, see Section 4.4.
IM 12E01A06-02E
4-4
<4. PARAMETER SETTING>
4.4
Service Level
The TB750G turbidimeter has various functions. At this level there are modes where functions can be
selected according to need for the operation.
Table 4.3
Code
CODE02
CODE08
CODE11
CODE12
CODE13
CODE14
CODE16
CODE17
CODE30
CODE33
Setting Parameters at Service Level (1/2)
Setting Parameter
Input voltage display
Transmitted light detection element input display
Scattered light detection element input display
End
Spike detection parameter setting
Spike detection function enabled or disabled
Limit value
Hold time
Sampling time
Zero calibration factor display
Zero calibration factor display
Slope display
Slope display
Zero correction factor setting
Zero correction factor setting
Sensitivity correction factor setting
Sensitivity correction factor setting
Standard solution/check block selection
Standard solution or check block
Check block turbidity setting
Check block turbidity
Analog output range
switching parameter setting
Range switching output selection
Range switching function selection
Range A setting
Zero point for range A
Span point for range A
Range B setting
Zero point for range B
Span point for range B
Range C setting
Zero point for range C
Span point for range C
Automatic range switching point
Analog output 2 range selection
Range selection for output 2
Fixed value for hold
Fixed value for hold during failure
CODE35
Hold during failure parameter setting
Hold function enabled or disabled
Last measured value, or fixed value
Fixed value for output 1
Fixed value for output 2
IM 12E01A06-02E
Display
Setting Range
Default
* IN1
* IN2
* END
* SPIKE
* LIMIT
* HLD-T
* SMP-T
0: Disabled, 1: Enabled
000.000 to 999.999
005 to 600[s]
001 to 600[s]
*CAL.A
* SL
*CAL.B –9.000 to 09.000[NTU]
*CAL.K 0.2500 to 4.0000
*CAL.TP 0: Standard Solution, 1: Check
Block
0:Disabled
999.999(NTU)
030(s)
030(s)
Depends on MS
Code
100.0(%)
00.000(NTU)
1.0000
1:Check Block
*PLATE 0.001 to 2000[NTU]
* RNGPR 0: Fixed, 1: AO1, 2: AO2
*RSET 0: Local, 1: Automatic,
2: Remote
2RNG, 3: Remote 3RNG
* RSET.A
* ZERO 0000.00 to 2000.00[NTU]
*SPAN 0000.00 to 2000.00[NTU]
* RSET.B
*ZERO 0000.00 to 2000.00[NTU]
* SPAN 0000.00 to 2000.00[NTU]
* RSET.C
* ZERO 0000.00 to 2000.00[NTU]
* SPAN 0000.00 to 2000.00[NTU]
*AUTOR 070 to 100[%]
*mA2 0: 4.0-20.0mA, 1: 0.0-20.0mA
0:Fixed
1:Automatic
0000.00(NTU)
0010.00(NTU)
0000.00(NTU)
0100.00(NTU)
0000.00(NTU)
1000.00(NTU)
080(%)
0:4.0-20.0mA
*H.mA2 00.0 to 22.0[mA](0-20mA)
02.0 to 22.0[mA](4-20mA)
*FH.mA2 00.0 to 22.0[mA](0-20mA)
02.0 to 22.0[mA](4-20mA)
22.0(mA)
* FHOLD 0: Disabled, 1: Enabled
*FH.LST / *FH.LST(Last measured
*FH.FIX value)
*FH.FIX(Fixed value)
*FH.mA1 02.0 to 22.0[mA]
*FH.mA2 00.0 to 22.0[mA](0-20mA)
02.0 to 22.0[mA](4-20mA)
1: Enabled
*FH.FIX:Fixed
value
22.0(mA)
22.0(mA)
22.0(mA)
4-5
<4. PARAMETER SETTING>
Table 4.3
Code
CODE37
CODE40
CODE41
CODE44
CODE50
CODE51
CODE52
CODE54
CODE61
CODE64
CODE66
CODE67
CODE71
CODE72
Setting Parameters at Service Level (2/2)
Setting Parameter
Time constant setting
Time constant in measurement
Time constant in maintenance
Contact output S1 function selection
Function of contact output S1
Contact output S2 function selection
Function of contact output S2
Delay time/hysteresis setting
Delay time
Hysteresis
Auto return function selection
Auto return function enabled or disabled
Alarm setting function selection
Alarm setting function enabled or disabled
Password setting
Password
Negative nondisplay/nonoutput selection
Negative nondisplay/nonoutput enabled or
disabled
Turbidity unit selection
Turbidity unit
Software version display
Software version display
Error “E2xx” error level selection
Error levels of errors “E201” to “E206”
Error “E3xx” error detection selection
Error detection for “E301” to “E306”
Error detection for “E307”
Error detection for “E311” to “E316”
Error detection for “E317”
Analog output test
Start test for output 1
Test current flowing
End
Start test for output 2
Test current flowing
End
Contact output test
Test 1
Test 2
CODE73
CODE79
Contact input test
Test
Initialization
Start
Default loading
Display
Setting Range
*TC 000 to 120[s]
*TC-M 000 to 120[s]
Default
020(s)
006(s)
* S1 0:None , 1: Alarm,
2: Automatic calibration/
washing, 3: Maintenance
1: Alarm
* S2 0:None , 1: Alarm,
2: Automatic calibration/
washing, 3: Maintenance
3: Maintenance
* D.TIME 000 to 199[s]
* HYST 000 to 100[%]
000(s)
002(%)
*RET 0: Disabled, 1: Enabled
0: Disabled
* MODE 0: Disabled, 1: Enabled
0: Disabled
*PASS 0:No password, 1:111, 2:333,
3:777,4:888, 5:123, 6:957,
7:331, 8:546, 9:847
*MINUS 0: Disabled, 1: Enabled
*UNIT 0: NTU, 1: FNU, 2: mg/l
*VER
0.0.0
0.0: Disabled.
Disabled
0: NTU
1.00
*201-6 0: Disable, 1: Level 1, 2: Level
2
1.1.2.2.2.2
*301-6
*307
*311-6
*317
1.1.1.1.1.1
1
1.1.1.1.1.1
1
0: Disabled, 1: Enabled
0: Disabled, 1: Enabled
0: Disabled, 1: Enabled
0: Disabled, 1: Enabled
*AO1.T
*END
*AO2.T
*END
*DO.T1 0: Non-action, 1: Action
*DO.T2 0: Non-action, 1: Action
Last measured
value held
Last measured
value held
*DI.T
*LOAD
*WAIT
IM 12E01A06-02E
4-6
<4. PARAMETER SETTING>
4.5
Procedures for Setting Parameters
An example for setting parameters, the procedures are described in following modes. For more detail
on the other functions, refer to User’s Manual IM 12E01A06-01E Section 6.
(1) CAL Mode: {CALIB}
(2) HOLD Mode: {HOLD}
(3) RANGE Mode: {*RANGE}
(1)
CAL Mode: {CALIB}
CALIB
[YES]
X.XXX
X.XXX
[NO]
STD.CAL
[YES]
SMP.CAL
[YES]
*1
*2
X.XXX
X.XXX
[NO]
ZERO
[NO]
SPAN
[YES]
*3
[NO]
*4 *5
0.000
VALUE
[ENT]
X.XXX
*5
[]
[]
[ENT]
X.XXX
ZERO
[NO]
ZRSPN
[YES]
0.000
VALUE
X.XXX
VALUE.Z
[]
[]
[ENT]
*5
[YES]
0.000
VALUE
ZERO
Self-adjustment
or [ENT]
*6
X.XXX
[NO]
SPAN
*5
[]
[]
[ENT]
X.XXX
SPAN
[NO]
ZERO
[YES]
0.000
X.XXX
X.XXX
ZERO
[YES]
0.000
VALUE
[]
[]
[ENT]
X.XXX
SPAN
X.XXX
SPAN
*5
[YES] *7
0.000
VALUE.S
[]
[]
[ENT]
Self-adjustment
or [ENT]
*6
[NO]
X.XXX
X.XXX
SPAN
Self-adjustment
or [ENT]
*6
Self-adjustment
or [ENT]
*6
Self-adjustment
or [ENT]
*6
Self-adjustment
or [ENT]
*6
CAL.END
[YES]
*7
HOLD
Measurement
mode
*1: During the time between [YES] key press on {STD.CAL} display and [YES] or [NO] key press on
{CAL.END} display or on error code display or [MODE] key press to abort, data display shows turbidity
based on sensitivity correction factor (K)=1 and zero correction factor (B)=0. (Value not affected by
SMP.CAL: Display by T1={(S0/(SL/100)}*(V-A), not by T2=K(T1+B))
*2: Span calibration in {STD.CAL} is performed using turbidity standards or check block that has been
selected in CODE16.
*3: In zero calibration, calibration value should be “0.000” and cannot be changed. Only [ENT] key is
accepted to confirm.
*4: In span calibration using check block, calibration value cannot be changed. Only [ENT] key is
accepted to confirm.
*5: First determine the decimal point using [>] key and then set value in the range of 0.000 to 20000 NTU.
*6: Pressing [ENT] key during self-adjustment will abort the procedure, and the turbidity when [ENT] key
is pressed will be used for calibration factor calculation.
*7: Only [YES] key is accepted to confirm.
Note: Calibration factor will be updated when the calibration procedure is completed without cancellation
or errors. If [YES]/[NO] key is pressed on error code display, the instrument goes to {CALIB} display.
Error E203 or E204 appears after the instrument once returns to measurement display.
F4-4E.ai
Figure 4.2
IM 12E01A06-02E
CAL Mode Flow Chart
<4. PARAMETER SETTING>
(2)
4-7
HOLD Mode: {HOLD}
In this mode whether the analog output is held or not in measurement mode, can be selected. This
function is available when the hold function is set to “1: Enabled” in SET HOLD mode at the setting
level. If the function is set to “0: Disabled,” this mode will be skipped.
HOLD
[YES]
(Flashing)
HOLD
[NO]
[YES]
Hold disabled in
measurement mode
Hold enabled in
measurement mode
HOLD status indicator turns on.
Measurement
mode
Figure 4.3
F4-5E.ai
HOLD Mode Flow Chart
If the hold function has been set to “1: Enabled” in SET HOLD mode at the setting level, the instrument
automatically goes into this HOLD mode when it returns from any level to measurement mode, with
the exception of DISPLAY mode.
(3)
RANGE Mode: {*RANGE}
In this mode, an output range corresponding to analog output 1 and 2 can be set. Enter turbidity
values corresponding to a 4 mA (or 0 mA) output signal (zero point) and a 20 mA output signal (span
point).
*RANGE
[YES]
*1
*FIXR.1
[NO]
*2
*FIXR.2
[YES]
XXXX.XX
*ZERO
[YES]
[]
[]
XXXX.XX
*ZERO
[YES]
XXXX.XX
*SPAN
*6
[ENT]
[NO]
[]
[]
[YES]
[]
[]
XXXX.XX
*SPAN
*6
[]
[]
*3
*LOCAL
[NO]
[YES]
*4
X
[]
Local range for output 1
X: 0(Range A), 1(Range B)
or 2(Range C)
[]
Local range for output 2
X: 0(Range A), 1(Range B)
or 2(Range C)
*LCL.1
[YES]
*5
X
*LCL.2
[ENT]
[ENT]
*1: Skipped if analog output range switching selection {*RNGPR} is set to “1: Analog output 1”
in CODE 30 at service level.
*2: Skipped if analog output range switching selection {*RNGPR} is set to “2: Analog output 2”
in CODE 30 at service level.
*3: Displayed only if analog output range switching selection {*RNGPR} is set to “1: Analog
output 1” or “2: Analog output 2” and range switching function selection {*RSET} is set to
“0: Manual” in CODE 30 at service level.
*4: Displayed only if analog output range switching selection {*RNGPR} is set to “1: Analog
output 1” in CODE 30 at service level.
*5: Displayed only if analog output range switching selection {*RNGPR} is set to “2: Analog
output 2” in CODE 30 at service level.
*6: Zero and span points should be set together. Only after span point has been entered, both
zero and span points are accepted. Cancellation by pressing [MODE] key during the
procedure will not change either of the points.
F4-6E.ai
Figure 4.4
RANGE Mode Flow Chart
IM 12E01A06-02E
Blank Page
5-1
<5. MAINTENANCE>
5. MAINTENANCE
5.1
Lamp Replacement
CAUTION
Wait for some time to replace lamp assembly because lamp temperature is high just after lamp power
was turned off.
Although the lamp has a normal service life of more than one year, it is recommended that the lamp
be replaced at an interval of one year in terms of preventive maintenance. Some lamps may fail within
one year.
The following describes how to replace the lamp.
(1) Remove power from the TB750G turbidimeter.
(2) Remove the cover of the light source unit on the left side of the detector by loosening the 4
setscrews.
(3) Disconnect the terminals for lamp from the terminal block. Remove the lamp assembly by
loosening the 2 setscrews.
(4) Replace the lamp assembly with a new one. Mount the new lamp assembly with the 2
setscrews. Connect the terminals for lamp to the terminal block (LAMP).
(5) Supply power to the TB750G turbidimeter.
(6) Check that the lamp is lit. Replace and fix the cover of the light source unit by tightening the 4
setscrews.
(7) After the lamp replacement, allow the turbidimeter to warm up for at least one hour. Then,
perform calibrations instructed in Sections 3.
(8) Lamp replacement is now complete.
Cover of the
Light Source Unit
Lamp Assembly
A
Cover for
Setscrews (2) Terminal
Terminal Block
Setscrews (2)
Terminal Block
Lamp Assembly
[View A]
Figure 5.1
F5-1E.ai
Lamp Replacement
IM 12E01A06-02E
5-2
<5. MAINTENANCE>
5.2 Calibration
5.2.1 Outline
(1)
CALIB
STD.CAL
SMP.CAL
ZERO
Zero calibration using zero turbidity
water
SPAN
Span calibration using check block
Span calibration using standard
solution
ZRSPN
2-point calibration using
standard solutions
ZERO
Zero point correction
SPAN
Sensitivity correction
(2)
(3)
Figure 5.2
Calibration
using
filtered
water as
zero
reference
F5-2E.ai
Calibration Methods
(1) Select when performing a calibration using 0.2 or 1.0 µm filtered water as zero reference.
(2) Select when performing a 2-point calibration using standard solutions. Use this method to
perform a calibration complying with EPA 180.1.
(3) Select when performing a zero point or sensitivity correction after calibration (1) or (2). This
method is used to adjust the meter reading to reflect the lab data in a grab sample calibration.
Note: The instrument has been pre-calibrated using 0.2 µm filtered water as zero reference at the
factory before shipment.
5.2.2 2-point Calibration Using Turbidity Standard Solutions
A 2-point calibration of the TB750G is performed with turbidity standard solutions equivalent to zero
solution and to span solution. After accepting both zero and span values, the meter calculates a
calibration factor.
When user-prepared zero and span solutions are used for calibrations, the turbidity of diluting water
for calibration should be measured with a calibrated lab turbidimeter in advance. Prepare a formazin
standard solution as a span solution. If the turbidity of diluting water is high against the measurement
accuracy, add the turbidity of diluting water to the nominal turbidity of the prepared formazin standard
solution. Use diluting water as zero solution and a formazin solution as span solution for the
calibration.
Refer to Section 3, “Operation” for concrete operating procedures.
IM 12E01A06-02E
6-1
<6. TROUBLESHOOTING>
6. TROUBLESHOOTING
Table 6.1
Error Code List (1/3)
No.
Error
Occurrence
Problem
E101
Flash memory
failure
When power is
turned on.
SUM results of flash memory do not
match programmed SUM values 5
consecutive times.
E102
EEPROM write
failure
All modes
Data is written in EEPROM and verified,
and write failure occurs 3 consecutive
times.
E103
RAM failure
When power is
turned on.
RAM area failure (3 consecutive times).
E104
AD converter
failure
All modes
AD converter failure (3 consecutive
times).
E201
Input voltage
failure
All modes
Either IN1 or IN2 input voltage is outside
the range of -0.15 to 1.2 V (fixed),
including abnormal value 0x7FFF or
8000, for 5 consecutive seconds.
E202
Detection
element failure
All modes
Either IN1 or IN2 input voltage is less
than specified voltage for 5 consecutive
seconds.
E203
Lamp life
expired
E204
Lamp intensity
failure
Zero calibration
coefficient A
rewriting timing.
(excluding when
E301 or E311
occurs.)*1
During zero calibration self-adjustment,
input voltage IN1 is less than specified
voltage for 5 consecutive seconds.
(Not detected when E301 is occurring.)
Rewriting of zero calibration coefficient A
is performed.
E205
Calibration
failure
E301 to E307,
E311 to E317,
E321 during PC
communication
occurrence
timing
Any of E301 to E307, E321 during PC
communication is occurring.
Return
When power is
turned off and then on
again, and problem is
eliminated. (Without
[YES]/[NO] key press.)
When problem is
eliminated for 5
consecutive seconds.
When [YES]/[NO] key
is pressed, error code
display disappears, at
the same time during
failure, error is once
removed.
When problem is
eliminated. When
[YES]/[NO] key is
pressed, error code
display disappears, at
the same time during
failure, error is once
removed.
IM 12E01A06-02E
6-2
<6. TROUBLESHOOTING>
Table 6.1
No.
Error Code List (2/3)
Error
Occurrence
Problem
Return
When problem is
eliminated. When
[YES]/[NO] key is
pressed, error code
display disappears, at
the same time during
failure, error is once
removed.
E301
Turbidity zero
calibration
coefficient A
failure
During zero
calibration (zero
water).
Turbidity zero calibration coefficient A
newly determined after zero calibration
(zero water or lamp OFF) is outside of
the range of 0 to 5, or each coordinate
value of newly determined linearized
table is outside of the range of -99999 to
999999.
E302
Turbidity slope
SL failure
During span
calibration
(standard
solution)
Turbidity slope value SL newly
determined after span calibration
(standard solution) is outside of the
range of 25 to 200%, or each coordinate
value of newly determined linearized
table is outside of the range of -99999 to
999999.
E303
Turbidity check
block failure
During span
calibration
(check block)
Turbidity slope value SL newly
determined after span calibration (check
block) is outside of the range of 50 to
150%, or each coordinate value of newly
determined linearized table is outside of
the range of -99999 to 999999.
E304
Turbidity zero
correction
factor B failure
During zero shift
calibration
Turbidity zero correction factor B newly
determined after zero shift calibration is
outside of the range of -9 to 9.
E305
Turbidity
sensitivity
correction
factor K failure
During span
calibration
(sensitivity
correction)
Turbidity sensitivity correction factor K
newly determined after span calibration
(sensitivity correction) is outside of the
range of 0.25 to 4.
E306
Turbidity
reference
sensitivity SO
failure
During turbidity
reference
sensitivity
calibration
Turbidity reference sensitivity SO newly
set after span calibration (check block)
is outside of the tolerance of 0.0001 to
2000, or each coordinate value of newly
determined linearized table is outside of
the range of -99999 to 999999.
E307
Turbidity
response time
failure
All calibrations
Turbidity self-adjustment is not complete
before self- adjustment time has
elapsed.
E321
Communication During
error
communication
with PC
IM 12E01A06-02E
Any communication error during
communication with PC.
When problem is
eliminated in the next
communication (only
one time is OK). When
[YES]/[NO] key is
pressed, error code
display disappears, at
the same time during
failure, error is once
removed. (excluding
RS signal during
communication with
calibration/cleaning
controller.)
6-3
<6. TROUBLESHOOTING>
Table 6.1
No.
Error Code List (3/3)
Error
Occurrence
Problem
Return
When [YES]/[NO] key
is pressed, error code
display disappears
and error is removed.
E351
Analog output
range setting
failure
During
parameter
setting in
RANGE mode or
CODE 30.
Either of:
(1) Zero point ≥ span point, or
(2) Span point - zero point < 20% of
span point or 0.20, whichever is greater
(for turbidity) or ( span point - zero point
< 5.00 (for color).
When automatic range is selected in
CODE 30, also possible:
(3) At each span point, Range A ≥ Range
B, or
(4) at each span point, Range B ≥
Range C
E352
Parameter
setting failure
During
parameter
setting
Value outside the setting range is set in
setting other than AO range setting.
Refer to the user’s manual IM 12E01A06-01E for details.
IM 12E01A06-02E
Blank Page
i
Revision Information
Model TB750G Right Angle Scattered Light Turbidimeter Quick Start Manual
Title:
Manual No.: IM 12E01A06-02E
Edition Date
Remark (s)
2nd
Mar. 2015
P.2-3,
P.6-1,
Some revision of Section 2.2.1, “Wiring Required for TB750G” :
Note 6 is added to figure 2.3.
Some revision of Section 6, “TROUBLESHOOTING” :
The name of error code E202 is revised, and “PD check voltage” and “lamp life check
voltage” changed to “specified voltage”.
1st
Newly published
Apr. 2006
IM 12E01A06-02E
Blank Page