Teledyne Analytical Instruments INSTRUCTION MANUAL MODEL 212R THERMAL CONDUCTIVITY ANALYZER
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Thermal Conductivity Analyzer
Model 212R
INSTRUCTION MANUAL
MODEL 212R
THERMAL CONDUCTIVITY
ANALYZER
DANGER
HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM.
PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.
HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST FOR A
TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.
ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING. BEFORE
CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/MANAGER.
Teledyne Analytical Instruments
P/N M73212
11/29/07
ECO # 07-0182
1
Thermal Conductivity Analyzer
Model 212R
All Rights Reserved. No part of this manual may be reproduced, transmitted,
transcribed, stored in a retrieval system, or translated into any other language or computer
language in whole or in part, in any form or by any means, whether it be electronic,
mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of
Teledyne Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 917491580.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us
free from defects of material and of construction, and that our liability shall be limited to
replacing or repairing at our factory (without charge, except for transportation), or at
customer plant at our option, any material or construction in which defects become
apparent within one year from the date of shipment, except in cases where quotations or
acknowledgments provide for a shorter period. Components manufactured by others bear
the warranty of their manufacturer. This warranty does not cover defects caused by wear,
accident, misuse, neglect or repairs other than those performed by Teledyne or an
authorized service center. We assume no liability for direct or indirect damages of any
kind and the purchaser by the acceptance of the equipment will assume all liability for
any damage which may result from its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our
apparatus, and to make any alterations in the dimensions, shape or weight of any parts, in
so far as such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its user, and serves as a tool by
which valuable data can be gathered. The information provided by the instrument may
assist the user in eliminating potential hazards caused by his process; however, it is
essential that all personnel involved in the use of the instrument or its interface, with the
process being measured, be properly trained in the process itself, as well as all instrumentation related to it.
The safety of personnel is ultimately the responsibility of those who control
process conditions. While this instrument may be able to provide early warning of
imminent danger, it has no control over process conditions, and it can be misused. In
particular, any alarm or control systems installed must be tested and understood, both as
to how they operate and as to how they can be defeated. Any safeguards required such as
locks, labels, or redundancy, must be provided by the user or specifically requested of
Teledyne at the time the order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions. The
purchaser is responsible for the training of personnel, for providing hazard warning
methods and instrumentation per the appropriate standards, and for ensuring that hazard
warning devices and instrumentation are maintained and operated properly.
Teledyne Analytical Instruments (TAI, the manufacturer of this instrument,
cannot accept responsibility for conditions beyond its knowledge and control. No
statement expressed or implied by this document or any information disseminated by the
manufacturer or its agents, is to be construed as a warranty of adequate safety control
under the user’s process conditions.
2
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Model 212R
Table of Contents
1
Introduction
1.1
1.2
1.3
1.4
1.5
2
Method of Analysis....................................................... 5
Sensitivity .................................................................... 6
Stability ........................................................................ 6
Special Consideration ................................................. 6
Physical Configuration ................................................. 6
Installation
2.1 Location ....................................................................... 7
2.2 Electrical Requirements & Connections ...................... 7
2.2.1 Primary Power ............................................... 8
2.2.2 Signal Output ................................................. 8
2.2.3 Regulating Transformer .................................. 8
2.2.4 Completion & Inspection ................................ 9
2.3 Gas Requirements & Connections ............................... 9
2.3.1 Reference Gas ............................................... 10
2.3.2 Zero Gas ........................................................ 10
2.3.3 Span Gas ....................................................... 10
2.3.4 Installation of Cylinder Supplies .................... 11
2.3.5 Sample Pressure ........................................... 11
2.3.6 Interconnecting Lines ..................................... 11
2.3.7 Vent Lines ...................................................... 12
3
Startup
3.1
3.2
3.3
3.4
3.5
3.6
3.7
3.8
4
Preliminary ................................................................... 12
Reference Gas Flow .................................................... 13
Zero Gas Flow ............................................................. 13
Warmup........................................................................ 13
Zero Standardization ................................................... 13
Span Standardization .................................................. 14
Bypass ......................................................................... 14
Sample Mode ............................................................... 15
Routine Operation
4.1 Flowrate ....................................................................... 15
4.2 Supporting Gas Supplies ............................................. 16
Teledyne Analytical Instruments
3
Thermal Conductivity Analyzer
Model 212R
4.3 Standardization ............................................................ 16
5
Periodic Maintenance
5.1 Heater Fan ................................................................... 17
6
Troubleshooting
6.1 Preliminary ................................................................... 17
6.1.1 Electrical Checks ........................................... 17
6.1.2 Sampling System Checks .............................. 18
6.1.3 Summary of Preliminary Checks .................... 18
6.2 Loss of Zero Control..................................................... 18
6.2.1 Dynamic Balance Procedure ......................... 19
6.3 Correct Operation ......................................................... 20
6.4 Incorrect Operation....................................................... 21
6.4.1 Analyzer Leak Check ..................................... 21
6.4.2 Temperature Control Check ........................... 22
7
Calibration Data
7.1
7.2
7.3
7.4
Ranges ........................................................................ 24
Output Signal ............................................................... 25
Span Setting ................................................................ 25
Recommended Accessory Gases ................................ 25
7.4.1 Reference Gas ............................................... 25
7.4.2 Zero Gas ........................................................ 25
7.4.3 Span Gas ....................................................... 26
Appendix
A
A
A
4
Recommended 2-Year Spare Parts List ....................... 26
Drawing List ................................................................. 27
Specifications...................................................................... 28
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
1.
1.1
Model 212R
INTRODUCTION
Method of Analysis.
The Model 212R compares the thermal conductivity of a sample gas to
that of a fixed composition reference gas and produces an electrical output
signal that is calibrated to represent the difference between the two gases.
Due to the nonspecific nature of thermal conductivity measurement,
standard gases of known composition will be required to calibrate the analyzer. The accuracy of the analysis will be dependent on the accuracy to
which the composition of the standard gases is known.
The measuring unit is a four element hot wire cell that forms one-half of
an alternating current bridge circuit. Two of the hot wire elements are exposed to the sample gas, and two to the reference gas. The other half of the
bridge circuit is formed by the center tapped secondary winding of a transformer.
With reference and zero standardization gas flowing, the bridge circuit
is balanced at one end of the measurement range. A span standardization gas
containing a known concentration of the component of interest is then
introduced into the sample path, and the resulting error signal generated by
the now unbalanced bridge circuit is calibrated to represent the span gas
mixture. The concentration of the component of interest in the span gas is
predicated by the specified ranges of the analysis. After the instrument has
been standardized the electrical error signal is directly related to the component of interest content of the sample gas.
The magnitude of the measuring bridge error signal is much too small to
drive an indicating or recording instrument. A 100:1 step-up transformer,
followed by an electronic amplifier stage, is utilized to amplify the error
signal to an acceptable amplitude for demodulation. The signal is then
conditioned appropriately to drive recording equipment.
Teledyne Analytical Instruments
5
Thermal Conductivity Analyzer
1.2
Model 212R
Sensitivity.
Differences in thermal conductivity that produce a measuring bridge error
signal of 0.25 microvolts will be sensed by the analyzer.
1.3
Stability.
Internal variables, other than the sample gas, that could produce variations
in the output signal are carefully controlled by the Model 212R. Zero drift is
approximately 2% of the fine range in a 24 hour period of operation; furthermore, the drift is bidirectional with a properly installed analyzer and is related to
gross changes in ambient temperature. The variables, both internal and external,
that can affect the stability of the analyzer will be dealt with in detail in the installation and trouble shooting sections of this manual.
1.4
Special Consideration.
Consideration of Using the 212R to Measure H2, or measure other compounds
in H2. H2 in the gas state can be assume of one tow states, para H2, or ortho
H2. Each has its own Thermal Conductivity Value. These differ by about 10%
from one another.
So this must be taken into consideration when attempting to use the 212R on H2
streams.
We would recommend one consult Wikipedia.com or other sources to learn
more about these two states of H2 and how the thermal conductivity of the gas
stream varies with the state.
If for instance one is using pure H2 as a reference gas, and this gas is in the para
state associated with cryogenic H2 recently vaporized, and this is then compared
to cylinder H2 which may be in a different state, then considerable measurement
errors can result.
Contact Teledyne for further guidance or information on your specific application.
1.5
Physical Configuration.
The analyzer is housed in a sheet steel case that is designed to flush mount
within an equipment panel. Electrical controls, as well as an integral gas control
6
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Model 212R
panel, are located immediately behind a hinged front access door. The analyzer
is suitable for installation in a sheltered non-hazardous area.
A recording and/or indicating device will be required to transduce the
electrical output signal into readable information.
2.
INSTALLATION
2.1
Location.
The analyzer should be panel mounted in an upright position in an area that
is not exposed to the following conditions:
1. Direct sunlight.
2. Drafts of air.
3. Shock and vibration.
4. Temperatures other than that one would expect to see in
airconditioned, temperature controlled office of lab enviroment.
The 212R should not be mounted outdoors or subject temperature fluctuations beyond 2 or 3 degrees.
The analyzer should be placed as close as possible, subject to the aforementioned conditions, to the sample point.
Outline diagrams of both units will be found in the drawing section. After
the cutout has been made in the equipment panel TAI recommends that the
analyzer be used as a template to lay out the four mounting holes. Such a procedure will compensate for sheet metal tolerance errors.
2.2
Electrical Requirements and Connections.
Provisions have been made to accommodate the three external circuit
connections required by the analyzer. Access holes on one side of the analyzer
case (see Outline Diagram) are provided for the installation of the conduit and
electrical wiring. All three customer connected circuits are to be terminated on
the barrier terminal strip identified “TS1”.
To install the conduit and wiring, the inner horizontally hinged panel must be
opened (to open the panel, turn the fastener screw at the top of the Panel a 1/4
turn ccw). While installing the conduit and wiring be careful not to disturb the
foam insulation lining the interior of the case anymore than is absolutely necessary.
Teledyne Analytical Instruments
7
Thermal Conductivity Analyzer
IMPORTANT:
Model 212R
The foam lining the interior of the analyzer is an
integral part of the environmental temperature
control system. Removal or destruction of this
lining will result in erratic instrument performance.
2.2.1 Primary Power.
A source of single phase, 105 to 125 volt, 60 cycle power will be required
to operate the analyzer. The maximum power consumption of the analyzer is
500 watts.
NOTE:
The analyzer is also available for 50 cycle operation
with special modifications and accessories.
Refer to the Interconnection Diagram in the drawing section of the manual
and connect the power and ground wiring as shown. Be sure to polarize the
power service connections as indicated. When connecting the wires, do not
leave an excessive amount of slack within the analyzer. Two vacuum tubes are
located just below the wiring area and the wiring should be installed to be well
clear of them.
2.2.2 Signal Output.
Connect a two conductor shielded cable between the analyzer and recording equipment. Be sure to observe the proper polarity at both instruments.
Connect the shield of the cable on the indicated terminal at the analyzer only, and
cut back and insulate the shield at the recorder.
NOTE:
Connecting the shield at both ends of a cable when
dealing with low level circuits can create a ground
loop between two instruments. Improperly installed
shielding can produce more noise in a low level circuit
than no shielding at all.
2.2.3 Regulating Transformer.
8
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Model 212R
Run the cable attached to the transformer unit in through the access hole
that is equipped with the cable clamp and terminate it as specified on the
interconnection diagram.
2.2.4. Completion and Inspection.
After the electrical connections have been completed, slide any excess
slack back into the conduits so that the installed wiring is not in contact with
the components mounted on the analyzer chassis.
Remove the three foam strips that are taped to the inside of the analyzer
door and use them to stuff the conduit openings. It is important that these
openings be as well sealed as possible.
Check to see that the temperature control printed circuit board, and all
vacuum tubes are firmly seated in their respective sockets.
Close and latch the control panel. There should be no further need to have
access to the interior of the analyzer. All controls and adjustments are arranged
so that they can be manipulated without disturbing the delicate temperature
equilibrium of the instrument interior.
2.3
Gas Requirements and Connections.
Before attempting installation of the sample and supporting gas lines and
accessories give careful consideration to the following important installation
notes.
Note #1: It is absolutely necessary that all connections and components
in the gas control system ahead of the measuring cell be leak
free. Toward that end TAI has tested the integral sampling
system under pressure with a sensitive leak detector and
certifies that the analyzer is leak free.
Note #2: Use no solder connections in the system. Soldering fluxes
outgas into the sample lines and produce erratic output
readings. Acid type soldering fluxes actually attack and
permanently change the characteristics of the detector cell
measuring elements.
Note #3: All sample system tubing should be new and clean. Many
gases and vapors are absorbed by dirt or oxide coatings on
tubing walls. These gases and vapors are released as the
ambient temperature rises. Because of the high sensitivity of
the analyzer, this absorption-desorption phenomenon can
Teledyne Analytical Instruments
9
Thermal Conductivity Analyzer
Model 212R
cause a large fluctuation in output signal. For best results, use
electropolished SS tubing
Note #4: Because of the diffusion of air through composition materials,
only dual stage SS regulators with metallic diaphragms should
be used in conjunction with the sample and supporting gas
supplies.
Note #5: Regulators must be thoroughly purged (burped) prior to use.
2.3.1 Reference Gas.
A cylinder of gas of fixed composition is required as the reference for
comparison with the sample gas. A total lack of impurities in the reference
would be ideal but is not necessary. An impurity concentration of up to 50%
of the narrowest range of interest is tolerable. The exact composition of the
reference gas is academic as long as the user is certain that it falls within the limits
specified in Section 7 of the manual. The important consideration is that the
composition of the reference gas remain unchanged when in use.
NOTE:
When it becomes necessary to replace the reference
gas, the analyzer will have to be recalibrated.
2.3.2 Zero Gas.
A supply of gas, composed of the background gas of the analysis and
the lowest attainable concentration of impurity, will be required to standardize one end of the ranges of interest. The composition of the zero gas must
be known to the same exactitude expected of the analysis. TAI suggests that
the gas be obtained from a supplier that will certify its composition. Recommendations as to the composition of the gas will be found in Section 7 of the
manual.
2.3.3 Span Gas.
A supply of gas, composed of the background of the analysis plus a
known concentration of the component of interest, will be required to standardize the sensitivity of the analyzer. Again, the composition of the gas
must be known to the same order of accuracy expected of the analyzer.
Analysis and certification of the composition is desirable. Specific recommendations governing the composition of the span gas will be found in
Section 7 of the manual.
10
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Model 212R
2.3.4 Installation of Cylinder Supplies.
The reference and standard gas cylinders should be installed as close to the
analyzer as possible. Each cylinder should be provided with a dual stage,
metallic diaphragm, pressure reducing regulator. When installing the regulators,
crack the cylinder valve open so that gas is flowing. This procedure will prevent
air from being entrapped in the regulator, and eliminate a common source of
supporting gas contamination. Improper installation of a cylinder regulator can
appreciably change the composition of a standard gas. Air trapped in the regulator will diffuse back into the cylinder, and the impurity concentration of the
composition will be altered. When dealing with the zero gas, the alteration of the
composition can be significant.
Once installed, prior to use, the regulators for reference, span and zero
gases must be purged by alteratively pressuring the first stage, with the second
stage off, then bleeding out the first stage by opening the second stage, back and
forth several times.
Note: Make sure to close the cylinder prior to bleeding the first stage
through the second.
2.3.5 Sample Pressure.
The sample point should be equipped with a metallic diaphragm pressure
regulator and the pressure reduced to between 10 and 50 psig. The regulator
should be installed as close to the sample point as possible to minimize sample
line lag time.
2.3.6 Interconnecting Lines.
The inlet and vent connections are identified on the analyzer outline diagram. The connections are stainless steel 1/8" female pipe couplings that are
braised into a gas connector bar mounted within the analyzer. The braised pipe
coupling was selected as a transition so that the torque generated during installation of the external system would be isolated from the internal system. Bulkhead
type fittings can be accidently twisted, and leaks promoted in hard to reach areas
within the analyzer. TAI suggests that 1/4" tubing and adapter fittings be used
throughout the external sampling and supporting gas system.
Teledyne Analytical Instruments
11
Thermal Conductivity Analyzer
Model 212R
2.3.7 Vent Lines.
The sample path, reference path, and an integral bypass system vent from
separate ports at the rear of the analyzer. Wherever possible, TAI recommends
that the gases be permitted to vent directly to the atmosphere. If it is necessary
to carry these gases to a remote area the following precautions must be observed when installing the vent lines:
1) The vent lines must be 1/4" tubing or larger so that no back pressure
resulting from restricted flow is experienced by the measuring cell.
2) The sample and reference paths must vent into an area that experiences the same ambient pressure conditions.
3) The ambient pressure at the vent location should undergo no more
than normal barometric pressure fluctuations.
4) The vent lines must be installed so that water and dust cannot accumulate in them.
A pressure differential existing at the cell between the reference and sample
will result in a corresponding change in output signal. The reference and sample
path flowmeters are located upstream from the cell so that both cell paths can
vent directly to atmosphere. The random bounce of the floats in the flowmeters,
when located downstream from the measuring cell, can produce up to 5% noise
on the output signal.
3.0
STARTUP
3.1
Preliminary.
The following preliminary steps should be accomplished before applying power or starting gas flow.
1. Check the integral gas control panel and be sure that all valves are
closed (fully cw). Do not jam the sample, reference, and bypass
metering valves.
2. Place the “RANGE” switch on the #2 position.
3. Set the “SPAN” control to the reading recorded in Section 7 of the
manual.
12
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Model 212R
4. Determine what the sample pressure is and set the output pressure of
the cylinder gas supplies to agree with the sample pressure. This
procedure, although not absolutely necessary, will minimize flow
adjustments when the sample path is switched to the various inputs.
5. Turn the analyzer and recorder power switches to “ON”.
3.2
Reference Gas Flow.
Open the reference gas throttle valve adjacent to the “Ref. Flow” meter,
and set the valve for an indicated flowrate of between 0.1 and 0.3 scfh. Since
the actual setting is not critical, TAI suggests that the flow be set at 0.1 scfh to
conserve the reference gas supply.
3.3
Zero Gas Flow.
Open the “ZERO” valve (ccw until a release in tension is felt), and adjust
the throttle valve to the right of the “SAMPLE FLOW’ flowmeter for an indicated flowrate of 0.3 scfh.
3.4
Warmup.
A warm-up period of twenty-four (24) hours is recommended to stabilize
the interior of the analyzer at the control temperature level. Until the measuring
cell temperature has equilibrated with the control temperature, the output signal
will drift.
During the warmup period, check the recorder periodically. As the drift
rate decreases, increase the sensitivity by moving the range switch towards
Range #1.
IMPORTANT:
The zero control will have little or no effect until Range #1 is reached. If the
recorder is off scale towards electrical zero (negative signal) after 24 hours of
running on zero gas, use the procedure outlined in Section 6.2. The drift rate will
decrease exponentially. When the rate of change slows to approximately 1%
per hour, place the “RANGE” switch on position #1 and readjust the “ZERO”
control so that the recorder is reading midscale. Do not attempt to calibrate the
analyzer until the recorder indication stabilizes.
3.5
Zero Standardization.
Teledyne Analytical Instruments
13
Thermal Conductivity Analyzer
Model 212R
After the analyzer stabilizes adjust the "ZERO" control until the recorder
indicates the impurity concentration of the zero gas composition. Be sure that
the range switch is on position #1 and that the sample path flowrate is 0.3 scfh.
NOTE if the"ZERO" or "FINE ZERO" pot runs out of adjustment and
the analyzer can not be calibrated, set this pot back to mid-scale. Then use
the "COURSE ZERO" pot to readjust the zero calibration as close as possible to zero calibration. And last use the "ZERO" ot "FINE ZERO" pot to
fine adjust the zero calibration.
3.6
Span Standardization.
After the zero has been standardized, switch the selector valve to "Span",
and reset the sample path flowrate to 0.3 scfh. The recorder should come to
balance on, or close to, the composition of the span gas.
The analyzer is factory calibrated to make sure that the output is linear over
all three ranges of interest. Calibration is achieved with mixing block technique
and the span setting recorded in Section 7 is derived at that time.
If the analyzer performed as described in the warm-up procedure (Section
3.4), and fails to closely approximate the composition of the span gas, there is
grounds for doubting the span gas mixture. Because of the difficulty involved in
obtaining precise analysis of small amounts of impurities in cylinder gas, and
because of the ease with which the gas can be contaminated subsequent to
analysis, any large error in response to the span gas should be suspect. In such a
case TAI recommends that the analyzer be operated at the recommended span
setting until the span gas is reanalyzed.
If adjustment of the span control is necessary to compensate for minor
discrepancies between the recorded setting and the span gas reading, TAI
suggests that the zero and span procedures be repeated until no further adjustment is required.
3.7
Bypass.
The integral gas control panel features a bypass flowmeter and throttle
valve that is located downstream from the input manifold (see the analyzer
piping schematic in the drawing section). The bypass system can be used to
speed the response of the analyzer to changes in the process. The bypass
flowmeter will indicate flowrates that are a factor of ten greater than the
sample path flowmeter. TAI recommends that the bypass system be used
whenever the sample path is switched and particularly after using the span
gas. The time required for the analyzer to stabilize on an impurity concentration within the limits of Range #1, after having been exposed to a concentration within the limits of Range #3, is not a function of cell response which is
14
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Model 212R
virtually instantaneous. Recovery is a function of how long it takes to purge the
sampling system upstream from the cell. Accelerating the sample path flowrate
will speed the response.
NOTE:
TAI recommends that some bypass flow be permitted
at all times. The bypass throttle meter is a part of the
input manifold. Being so located means that the
throttle valve lies between the sample side of the cell
and atmosphere.
Without some bypass flow it is possible that atmosphere could diffuse into the sample line. Do not
attempt to use the throttle valve as a shut off valve.
Jamming the valve will damage the metering needle.
3.8 Sample Mode.
After the analyzer has been standardized, switch the selector valve to
"Sample". Set the bypass flow so that the float is visibly elevated from the
bottom of the tube and then readjust the sample flowrate to 0.3 scfh. No
further adjustments are required.
NOTE:
Although the bellows type valves used in the input
manifold are among the best available, TAI suggests
that the span gas be turned OFF at the cylinder when
not in use, and the span line relieved of pressure
before closing the analyzer span valve. Such a procedure will eliminate any possibility of the span gas
diffusing into the sample path of the analyzer. Normally, the recommended span gas composition contains a component of interest content of 3/4 the coarsest range of analysis. The component of interest
content of the sample gas, on the other hand, is usually well within the finest range of analysis. Instrument
sensitivity in Range #1 is 100 times greater than in
Range #3. The user can readily see what even a
minute rate of span gas diffusion would do to the
validity of a Range #1 analysis.
4.
4.1
ROUTINE OPERATION
Flowrates.
Teledyne Analytical Instruments
15
Thermal Conductivity Analyzer
Model 212R
The reference and sample flowrates should be checked daily. The instrument is somewhat flow sensitive. Operating on Range #1 a change in flowrate of
0.1 scfh will cause a corresponding change in sensitivity of from 1 to 20% of
scale. On the coarser ranges, the change in sensitivity would be undetectable.
The sample path flowrate should be maintained at 0.3 scfh, the reference
path flowrate at 0.1 scfh, and the bypass flowmeter float should be slightly above
the bottom of the indicator glass.
4.2
Supporting Gas Supplies.
Supporting cylinder gas supplies should be checked frequently on a routine
basis with particular attention focused on the reference gas. A spare cylinder of
reference gas should be available at all times. When the cylinder pressure drops
below 100 psig the reference supply should be replaced as the operation of the
regulator is questionable at pressures below this point.
When replacing supply cylinders, be sure to bleed the gas through the
cylinder valve while installing the pressure regulator (see Section 2.3.4). It is
also advisable to check the connections with soap water whenever a supply
cylinder is changed.
4.3
Standardization.
The analyzer must be completely restandardized whenever the reference
gas supply is replaced. Barring unforeseen difficulties with the analyzer, restandardization should not be necessary between reference cylinder replacement
periods if the analyzer is run continuously. If the analyzer is shut down for long
period of time use the startup procedure (in its entirety) when operation is to be
resumed.
NOTE:
16
TAI strongly recommends that the analyzer run continuously with gas flowing in both the sample and
reference paths. During inactive periods arrange the
input manifold so that zero gas is flowing. Gas supplies can be conserved by reducing the sample and
reference path flowrates to less than 0.1 scfh and
closing off the bypass flow completely.
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Model 212R
Due to the delicate balance of the thermal conductivity cell employed in the
analyzer, exposure to the atmosphere in an uncontrolled temperature environment is undesirable. After extended shutdown periods many days of operation
may be required for the analyzer to re-stabilize.
5.
PERIODIC MAINTENANCE
5.1
Heater Fan.
The heater fan is the only component within the analyzer that requires
periodic attention. Whenever the reference gas is replaced the motor bearings will require a few drops of light machine oil. Since oiling the fan motor
necessitates opening the inner door of the analyzer, the instrument will require the
necessary temperature equilibrating time after the door is closed. With zero gas
flowing be sure that the analyzer has stabilized on the Range #1 position before
re-standardization is attempted.
6.
6.1
TROUBLE SHOOTING
Preliminary.
If the analyzer is suspected of incorrect operation, as a preliminary to
evaluation, always arrange the input manifold so that zero gas is flowing
through the sample path of the analyzer. Never attempt to evaluate the
performance of the instrument with sample gas flowing.
Analysis by thermal conductivity is nonspecific in nature. A thermal conductivity analyzer with the fine range sensitivity of the Model 212R will respond
to many influencing factors other than the component of interest; particularly
when operated at Range #1 sensitivity.
It is necessary, therefore, to eliminate as many external variables as
possible if the performance of the analyzer is to be assessed. Programming
the analyzer for zero gas places both paths of the measuring cell on relatively
reliable cylinder gas sources.
6.1.1
Electrical Checks.
Teledyne Analytical Instruments
17
Thermal Conductivity Analyzer
Model 212R
Check to see that both the analyzer and recorder are being furnished
electrical power. Check to see that the analyzer and recorder fuses are intact.
Check all electrical connections outside of the analyzer.
6.1.2
Sampling System Checks.
Be sure that there is an adequate supply of reference gas and that the
sample and reference path flowrates are correct. Check to see that there are
no obstructions in the vent paths from the analyzer. Check all external
plumbing connections for leaks with soap water.
6.1.3
Summary of Preliminary Checks.
The analyzer inner door should not be opened and no adjustments, beyond
manipulation of the normally used controls, should be made until all the aforementioned preliminaries have been completed and any necessary remedial
repairs effected.
6.2
Loss of Zero Control.
If loss of zero control on Range #1 is experienced during initial startup or
after the reference gas supply has been changed, the dynamic balance procedure
must be repeated.
Loss of zero control, under such circumstances, indicates that the impurity
concentration within the reference gas is different than that of the gas used during
factory checkout (or the previous cylinder) and does not in itself indicate a
defective analyzer.
The analyzer was adjusted at the factory to have a zero balance at close to
the midpoint of the zero control potentiometer (dual dial reading of 500) with a
common source of high purity cylinder gas supplying both the sample and
reference paths of the cell. If the recorder is off scale and cannot be returned
with the zero control, the dynamic balance procedure must be employed to
restore control before further conclusions as to instrument performance can be
made.
18
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
6.2.1
Model 212R
Dynamic Balance Procedure.
The measuring bridge circuit (see schematic) incorporates two 10 turn
potentiometers that are utilized to compensate for minute differences in the
resistive and reactive properties of the hot wire elements and the center
tapped secondary of the bridge transformer in the reference and zero gas
related to the application.
The potentiometers are of the screw driver adjustment type and are
equipped with shaft locking assemblies that have been factory adjusted to create
enough friction so that the shaft will not change position in shipment. The potentiometers are located on the amplifier chassis, and are controlled through two
holes identified “RESISTIVE” and “REACTIVE” on the inner control panel.
DO NOT OPEN THE INNER PANEL OF THE ANALYZER TO
GAIN ACCESS TO THESE CONTROLS. The balance adjustments
should not be disturbed if the analyzer is not up to operating temperature.
Correct bridge balance can only be achieved at operating temperature because of the inherent thermal junctions within the potentiometers themselves.
The minute voltages produced by contact between the dissimilar materials
within the potentiometers are effectively balanced out in the procedure when
it is accomplished at operating temperature. Because the voltage generated
by these junctions varies with temperature, and because the analyzer will
resolve a 0.25 microvolt change in bridge voltage, temperature stability is a
mandatory prerequisite to the dynamic balance procedure.
To restore zero control within the limits of Range # 1, employ the
following Procedure:
1) With zero gas flowing, adjust the sample and reference gas flowmeters to the same value between 0.1 and 0.3 scfh.
2) Set the ZERO potentiometer to 500.
3) Connect a voltmeter to the instrument signal output. Connect an
oscilloscope to the probe point an d set the sweep rate to 5
milliseconds. Set the voltage level so ± 15 volts can be read.
4) If a reading of +15 or –15 volts is indicated on the oscilloscope
with the RANGE switch set to Range # 1 the instrument is
significantly unbalanced. Set to Range #2 and proceed to step 4).
If the voltage is lower in magnitude than + or – 15 skip down to
step 6).
Teledyne Analytical Instruments
19
Thermal Conductivity Analyzer
Model 212R
5) Using a screwdriver that has a shaft diameter of 1/8” and a shaft
length of at least 6”, turn the “Reactive” balance adjustment all
the way counterclockwise (CCW) until the potentiometer stop
point is reached. Turn the adjustment back clockwise (CW) 5 full
turns to center the adjustment.
6) Turn the “Resistive” balance adjustment all the way clockwise (CW)
until the potentiometer stop point is reached.
7) Slowly turn the “Resistive” adjustment until the output reaches its
lowest value. If the reading is negative adjust to the most negative value possible.
8) Turn the “Reactive” balance adjustment slowly CW and CCW to
obtain the lowest possible output value. Again, if the reading is
negative adjust to the most negative value possible.
9) Repeat steps 6) and 7) until the output value is as low as these
adjustments can make it.
10) Detune the instrument by turning the “Resistive” balance adjust
ment ½ turn CW. This will bring the signal out of the baseline
noise, which could otherwise cause non-linear response at the
low concentration end.
NOTE:
Due to minute differences between measuring cells
and the electronics in the bridge circuit found in various instruments, some analyzers will have to be balanced at a point higher than 10% of scale while others
can be balanced even closer to electrical zero. The
10% of scale recommended in step 6 of the procedure
is an arbitrary starting point. If the bridge circuit can
be easily balanced at this signal level, the procedure
should be repeated at a point even closer to electrical
zero. Conversely, if balance cannot be accomplished
at 10% of scale, the operator will have to select a
higher recorder set point for the procedure. In any
case, the object is to accomplish dynamic balance
with the lowest magnitude of analyzer cutout signal
attainable with the analyzer at Range #1 sensitivity.
6.3
Correct Operation.
If a stable recording is achieved when the analyzer is operated on zero
gas, erratic performance must be attributed to the customers sampling system
20
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Model 212R
or incorrect instrument installation. The following conditions can produce an
erratic output from the analyzer:
1) Installation in an area that is subject to large changes in ambient
temperature, direct sunlight, drafts from wind, blowers, or air conditioners. Any of the foregoing conditions will produce a bidirectional
(diurnal) variation in output signal.
2) Sample gas with a moisture content of 10 ppm or greater. Again, the
output will vary bi-directionality with ambient temperature changes.
If the sample gas is known to have a fairly high moisture content a
dryer should be installed in the sample line.
3) Even the smallest leak anywhere in the sampling system upstream
from the analyzer, a small leak at a fitting, valve, or regulator, will
cause almost constant variation in analyzer output. The resulting
signal drift would be erratic and could be either bi-directional or
unidirectional.
4) Acid in the sampling system (soldering fluxes) will produce continuous, uncontrollable, unidirectional output signal drift. Sample tubing
that has been treated with descaling chemicals, if subject to above
ambient temperatures, will outgas continuously. Any form of acid in
the sampling system will outgas at high temperatures and attack the
hot wire elements of the cell.
5) Composition materials (diaphragms, tubing, etc.) in the sampling
system will produce erratic, unpredictable changes in output signal
level because of their permeability to gases other than the sample.
The nylon tubing normally used in quick disconnect type sampling
systems will produce diurnal output signal excursions in the order of
10 to 20% of the fine range of analysis.
6.4
Incorrect Operation.
If the analyzer is not stable on zero gas, and the supporting gas installation
has been made and checked in accordance with sections 2 and 3, the following
procedures should be employed.
6.4.1 Analyzer Leak
Teledyne Analytical Instruments
21
Thermal Conductivity Analyzer
Model 212R
Check as previously stated, the analyzers entire sampling system has been
checked under pressure with a leak detector. However, if the external system
checks out and the instrument is displaying a large diurnal or uni- directional drift,
the interior sampling system should be checked for leaks. Do not open the
analyzer inner door. Use the following procedure:
1)
2)
Acquire a source of cylinder helium and through a dual stage
regulator connect it to a length of 1/8" tubing.
Open the outer door of the analyzer and insert the tubing between the edge of the sample control panel and the outer case.
CAUTION: Do not insert the tubing along the edges of the
larger swing-down control panel as there is danger of
obstructing the fan or short circuiting the electronics.
The end of the tubing should be shaped into a curve
so that the foam lining is not overly disturbed and the
tubing should only be inserted a couple of inches into
the interior.
3)
Start the helium flowing. Keep the flowrate low so as not to
disturb the temperature equilibrium of the analyzer interior.
4)
Observe the recorder. If any leaks are present in the interior
sampling system, the recorder will be driven off scale in one
direction or the other. NOTE: The analyzer should be on Range
#1 and the zero control adjusted so that the recorder is reading
mid-scale. The direction the recorder indication moves is a
function of the type of zero and reference gas being used (which
varies per application) and which path of the sampling system
(sample or reference) the leak is located in. If the recorder
responds to an atmosphere of helium within the analyzer, all
connections inside the analyzer must be checked.
6.4.2 Temperature Control Check.
The interior temperature of the analyzer is regulated to a constant value to
within .002 degrees centigrade. The actual temperature control point varies
slightly per application and is a function of the components used in the proportional temperature control circuit.
If the recorder is displaying a large diurnal recording that is related to the
time of day in terms of the differential between day and night ambient temperature, trouble in the temperature control circuitry is indicated. Again, this symp22
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Model 212R
tom must be evident with zero gas flowing. If this symptom is displayed only
when the sample is flowing, the problem is related to one (or more) of the five
(5) points outlined in section 7.3.
To check the temperature control circuit, use the following procedures:
1)
With the power switch ON, open the control panel door and
check to see if the fan is running. If it is not, check the left hand
fuse. This fuse protects the entire temperature control circuit
(see schematic). If the fuse will not hold, a short circuit is
indicated. Disconnect the printed circuit card and replace the
fuse. If the fuse holds (as indicated by the fan running) the
printed circuit card will have to be replaced.
NOTE: Unless competent electronic technicians are available, TAI
recommends that a replacement printed circuit card be ordered
and the existing board returned to the factory for repair. Repair
charges (out of warrantee) will be based on time and material.
A schematic of the proportional temperature control card is
included among the drawings at the rear of the manual.
2)
If the fuse will not hold with the board removed, a short is indi
cated in the fan, the heater resistors, or the interconnecting
wiring. With the board removed, check wiring and components
for short circuits.
3)
If the fan is running, connect an AC volt meter across anyone of
the heater resistors on the fan assembly (the meter should be set
to read 50 VAC). In a properly operating circuit, the meter
should read approximately 47 volts (heater control full on) as the
control panel is open and the thermistor should be demanding
maximum heater voltage. If no voltage is present, either the
circuit card or the thermistor probe is faulty. Isolate the problem
by first disconnecting the circuit board from its socket and with
an ohmmeter check between terminals 5 and 6 of terminal strip
TS2. If a reading is obtained (disregard actual resistance) the
thermistor is intact and a new circuit card will be necessary. If
no reading is obtained, a new probe assembly will be required.
4)
If approximately 47 VAC is measured across the heater resistors, a runaway, or correct, circuit is indicated. To determine
which is the case, hold a soldering iron in close proximity (1/2")
to the end of the thermistor probe and observe the meter.
Teledyne Analytical Instruments
23
Thermal Conductivity Analyzer
Model 212R
CAUTION: Do not touch the thermistor probe with the iron. The
end of the probe is tied and taped to the sample tubing
below the cell cover. If the meter reading suddenly
drops to zero, remove the soldering iron immediately.
After the area around the thermistor cools, the voltage
across the heater resistors should again climb to 47
volts. Such action indicates a properly functioning
heater circuit. If the meter stays in the 47 volt region,
the circuit card will have to be replaced.
7.0
CALIBRATION DATA FOR MODEL 212R
SERIAL NUMBER: 195929
7.1
Ranges.
The ranges of the analyzer are:
24
Range Switch Position #1:
0-50,000 ppm N2 in H2
Range Switch Position #2:
0-5000 ppm N2 in H2
Range Switch Position #3:
0-500 ppm N2 in H2
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
7.2
Model 212R
Output Signal:
Output coincides with 100% analysis.
4mA
=
0 ppm N2 in H2
20mA
=
50,000 ppm N2 in H2 - Range #1
=
5000 ppm
N2 in H2 - Range #2
=
500 ppm
N2 in H2 - Range #3
7.3
7.4
H2 Purity
H2 Purity
Span Setting: 555
Recommended Accessory Gases.
Zero Gas
Reference Gas
Span Gas
99.999
99.999
350-450 ppm N2 in H2
7.4.1 Reference Gas.
With a equivalent impurity of less than 3% of range 1. The impurity ratio
must remain constant. When the reference gas supply is replaced, the analyzer
must be re-standardized.
7.4.2 Zero Gas.
Because of the difficulties involved in obtaining a gas in pure form, the zero
gas must be purchased from a supplier who will certify the cylinder as to its
content. The equivalent impurity concentration of the gas must be less than 3%
of range 1. When the zero gas is introduced for standardization, the range
switch must be in Range 1, and the zero control set such that the recorder
indicates the zero gas impurity.
IMPORTANT: At startup, be sure that the span control dial is set to the
number recorded in Section 7.3.
Teledyne Analytical Instruments
25
Thermal Conductivity Analyzer
7.4.3
Model 212R
Span Gas.
Containing an equivalent impurity. Again, the span gas must be procured
from a supplier who will certify its composition. The range switch must be in
range 3 when the span gas is introduced.
IMPORTANT: The accuracy of the analysis is directly related to the
users knowledge of the span gas composition.
Reference and Sample Flowrates 0.3 SCFH
Bypass Flowrate 3.0 SCFH
NOTE:
Cylinder and sample stream pressure settings must
be adjusted in order to maintain the above flowrates
when switching from one sample to another.
RECOMMENDED SPARE PARTS LIST
QUANTITY PART NO.
1
A-6251
1
A-6981
1
F-45
1
F-22
DESCRIPTION
Measuring Cell Block Assembly
Proportional Temperature Control
Sample Flowmeter Tube
Sample Flowmeter Bypass Tube
A minimum charge applies to all spare part orders.
IMPORTANT: Orders for replacement parts should include the part
number, the model and serial number of the analyzer in which they are to be
used.
26
Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Model 212R
Orders should be sent to:
TELEDYNE Analytical Instruments
16830 Chestnut Street
City of Industry, CA 91749-1580
Phone (626) 934-1500, Fax (626) 961-2538
TWX (910) 584-1887 TDYANYL COID
Web:
www.teledyne-ai.com or your local representative.
Drawing List
B-73356
D-73212
A-72035
C-73085
C-73088
Outline Diagram
Final Assembly
Piping Diagram
Main PCB Schematic
Power Supply PCB Schematic
Replacement Parts Order Information
A minimum charge of $150.00 is application to all spare parts
orders.
Important: Orders for replacement parts should include the art
number (if available) and the model and serial number of the instrument
for which the part is intended.
Teledyne Analytical Instruments
16830 Chestnut street
City of Industry, Ca 91748-1580
Telephone: (626) 961-2538
(626) 934-1500
Fax:
(626) 961-2538
(626) 934-1651
Teledyne Analytical Instruments
27
Thermal Conductivity Analyzer
Model 212R
APPENDIX
Specifications:
Ranges: Three decade ranges: Minimum Full scale(ppm)
20 ppm Hydrogen balance Argon
25 ppm Hydrogen balance Nitrogen, Air
100 ppm Nitrogen balance Argon, Helium
100 ppm Helium balance Air
150 ppm Nitrogen, Carbon Monoxide balance Hydrogen
200 ppm Oxygen balance Argon
Range ID:
Detector:
3 range ID contacts
Local Range ID LED’S
Thermal conductivity sensor
Signal Output:
0-1VDC negative ground
4-20maDC Isolated ground
Accuracy:
Operating
Temperature:
Sample
Requirement:
Display:
Mounting:
Response Time:
Stability:
Power
requirement:
28
±2% full scale for most binary mixture
at constant temperature. ± 5% of Full
Scale over operating temperature (once
temperature equilibrium has been achieved)
20 – 30°C (68° F to 85° F)
Sample: 0.3SCFH
Reference: 0.1SCFH
Digital LED read out (3½ Digit)
Flush panel mount.
90% of full scale less then 60 seconds.
For most application.
Less then 2% full scale drift over 24
hours
110VAC, 50/60Hz (220VAC optional)
Teledyne Analytical Instruments