Endress + Hauser Operating Instructions Turbimax CUE21 / CUE22 Turbidimeter for online measurement
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Products
Solutions
Operating Instructions
Turbimax CUE21 / CUE22
Turbidimeter for online measurement
MODE
EXIT
OUT
BA00395C/07/EN/13.15
71292787
OUT
Turbimax CUE21
IN
E
Services
Table of contents
Turbimax CUE21 / CUE22
Table of contents
1
Document information . . . . . . . . . . . . . . 3
1.1
1.2
Safety information . . . . . . . . . . . . . . . . . . . . . . 3
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2
Basic safety instructions . . . . . . . . . . . . 4
2.1
2.2
2.3
2.4
2.5
Requirements for personnel . . . . . . . . . . . . . . .
Designated use . . . . . . . . . . . . . . . . . . . . . . . .
Workplace safety . . . . . . . . . . . . . . . . . . . . . . .
Operational safety . . . . . . . . . . . . . . . . . . . . . .
Product safety . . . . . . . . . . . . . . . . . . . . . . . . .
3
Incoming acceptance and product
identification . . . . . . . . . . . . . . . . . . . . . . . 6
3.1
3.2
3.3
3.4
Incoming acceptance . . . . . . . . . . . . . . . . . . . .
Product identification . . . . . . . . . . . . . . . . . . . .
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . .
Certificates and approvals . . . . . . . . . . . . . . . .
4
Installation . . . . . . . . . . . . . . . . . . . . . . . . 8
4.1
4.2
4.3
Installation requirements . . . . . . . . . . . . . . . . . 8
Mounting the measuring device . . . . . . . . . . . . 9
Post-installation check . . . . . . . . . . . . . . . . . . 12
5
Electrical connection . . . . . . . . . . . . . . 13
5.1
5.2
5.3
5.4
5.5
5.6
Connection requirements . . . . . . . . . . . . . . . .
Connecting RS-485 cables . . . . . . . . . . . . . . .
Connecting relays . . . . . . . . . . . . . . . . . . . . .
Connecting the current output . . . . . . . . . . . .
Ensuring the degree of protection . . . . . . . . . .
Post-connection check . . . . . . . . . . . . . . . . . .
6
Operation options . . . . . . . . . . . . . . . . . 16
6.1
6.2
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Operating concept . . . . . . . . . . . . . . . . . . . . . 16
7
Commissioning . . . . . . . . . . . . . . . . . . . . 18
7.1
7.2
Function check . . . . . . . . . . . . . . . . . . . . . . . 18
Switching on the measuring device . . . . . . . . . 18
8
Operation . . . . . . . . . . . . . . . . . . . . . . . . . 19
8.1
8.2
Configuring the measuring device . . . . . . . . . . 19
Instrument calibration . . . . . . . . . . . . . . . . . . 23
9
Diagnostics and troubleshooting . . . 27
9.1
9.2
General troubleshooting . . . . . . . . . . . . . . . . . 27
Diagnostic information on local display . . . . . . 27
10
Maintenance . . . . . . . . . . . . . . . . . . . . . . 29
10.1
Cleaning the flow-through cuvette . . . . . . . . . 29
2
4
4
4
5
5
6
6
7
7
10.2
10.3
Replacing the flow-through cuvette . . . . . . . . 29
Replacing the desiccant pouch . . . . . . . . . . . . 30
11
Repairs . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
11.1 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . .
11.2 Replacing the electronics module . . . . . . . . . .
11.3 Replacing the lamp . . . . . . . . . . . . . . . . . . . .
11.4 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
11.5 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . .
12
31
31
31
32
32
Accessories . . . . . . . . . . . . . . . . . . . . . . . 33
12.1 Calibration solutions . . . . . . . . . . . . . . . . . . . 33
12.2 Flow chamber . . . . . . . . . . . . . . . . . . . . . . . . 33
12.3 Passive barrier . . . . . . . . . . . . . . . . . . . . . . . . 33
13
Technical data . . . . . . . . . . . . . . . . . . . . 34
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
13
14
14
14
14
14
Endress+Hauser
Turbimax CUE21 / CUE22
Document information
1
Document information
1.1
Safety information
Structure of information
LDANGER
Causes (/consequences)
Consequences of non-compliance
(if applicable)
‣ Corrective action
LWARNING
Causes (/consequences)
Consequences of non-compliance
(if applicable)
‣ Corrective action
LCAUTION
Causes (/consequences)
Consequences of non-compliance
(if applicable)
‣ Corrective action
NOTICE
Cause/situation
Consequences of non-compliance
(if applicable)
‣ Action/note
1.2
Symbol
Meaning
This symbol alerts you to a dangerous situation.
Failure to avoid the dangerous situation will result in a fatal or serious
injury.
This symbol alerts you to a dangerous situation.
Failure to avoid the dangerous situation can result in a fatal or serious
injury.
This symbol alerts you to a dangerous situation.
Failure to avoid this situation can result in minor or more serious injuries.
This symbol alerts you to situations which may result in damage to
property.
Symbols
Meaning
Additional information, tips
Permitted or recommended
Not permitted or not recommended
Reference to device documentation
Reference to page
Reference to graphic
Endress+Hauser
Result of a step
3
Basic safety instructions
Turbimax CUE21 / CUE22
2
Basic safety instructions
2.1
Requirements for personnel
• Installation, commissioning, operation and maintenance of the measuring system may
be carried out only by specially trained technical personnel.
• The technical personnel must be authorized by the plant operator to carry out the
specified activities.
• The electrical connection may be performed only by an electrical technician.
• The technical personnel must have read and understood these Operating Instructions
and must follow the instructions contained therein.
• Measuring point faults may be repaired only by authorized and specially trained
personnel.
Repairs not described in the Operating Instructions provided may only be carried out
directly by the manufacturer or by the service organization.
2.2
Designated use
The turbidimeters from the Turbimax CUE21 / CUE22 series are designed for online
measurement of turbidity in process water and drinking water.
The infrared device CUE21 meets the design criteria for turbidity measurement specified
by ISO 7027 and DIN 27027.
The white light device CUE22 meets the design criteria specified by US EPA 180.1.
Use of the device for any purpose other than that described, poses a threat to the safety of
people and of the entire measuring system and is therefore not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.
2.3
Workplace safety
As the user, you are responsible for complying with the following safety conditions:
• Installation guidelines
• Local standards and regulations
Electromagnetic compatibility
• The product has been tested for electromagnetic compatibility in accordance with the
applicable European standards for industrial applications.
• The electromagnetic compatibility indicated applies only to a product that has been
connected in accordance with these Operating Instructions.
4
Endress+Hauser
Turbimax CUE21 / CUE22
Basic safety instructions
2.4
Operational safety
1.
Before commissioning the entire measuring point, verify that all connections are
correct. Ensure that electrical cables and hose connections are undamaged.
2.
Do not operate damaged products, and safeguard them to ensure that they are not
operated inadvertently. Label the damaged product as defective.
3.
If faults cannot be rectified:
Take the products out of operation and safeguard them to ensure that they are not
operated inadvertently.
2.5
Product safety
The product is designed to meet state-of-the-art safety requirements, has been tested, and
left the factory in a condition in which it is safe to operate. The relevant regulations and
European standards have been observed.
Endress+Hauser
5
Incoming acceptance and product identification
Turbimax CUE21 / CUE22
3
Incoming acceptance and product
identification
3.1
Incoming acceptance
1.
Verify that the packaging is undamaged.
Notify your supplier of any damage to the packaging.
Keep the damaged packaging until the matter has been settled.
2.
Verify that the contents are undamaged.
Notify your supplier of any damage to the delivery contents.
Keep the damaged products until the matter has been settled.
3.
Check the delivery for completeness.
Check it against the delivery papers and your order.
4.
Pack the product for storage and transportation in such a way that it is protected
against impact and moisture.
The original packaging offers the best protection.
The permitted ambient conditions must be observed (see "Technical data").
If you have any questions, please contact your supplier or your local sales center.
3.2
Product identification
3.2.1
Nameplate
The nameplate provides you with the following information on your device:
• Manufacturer identification
• Order code
• Extended order code
• Serial number
• Ambient and process conditions
• Safety information and warnings
Compare the data on the nameplate with your order.
3.2.2
Product identification
Product page
www.endress.com/cue21
www.endress.com/cue22
Interpreting the order code
The order code and serial number of your product can be found in the following locations:
• On the nameplate
• In the delivery papers
Obtaining information on the product
6
1.
Go to the product page for your product on the Internet.
2.
In the navigation area on the right-hand side, select "Check your device features"
under "Device support".
An additional window opens.
Endress+Hauser
Turbimax CUE21 / CUE22
Incoming acceptance and product identification
3.
Enter the order code from the nameplate into the search field.
You will receive information on each feature (selected option) of the order code.
3.3
Scope of delivery
The scope of delivery comprises:
• 1 turbidimeter Turbimax CUE21 or CUE22
• 1 field terminal box
• 1 packet of desiccant
• 1 tubing kit including
– 1 shutoff clamp
– 1 flow controller
– 2 connecting tubes with fittings for flow-through assembly
– 1 vent screw for drain tube (used in pressurized systems)
• 1 Operating Instructions BA00395C
If you have any questions, please contact your supplier or your local sales center.
3.4
Certificates and approvals
The product meets the requirements of the harmonized European standards. As such, it
complies with the legal specifications of the EC directives. The manufacturer confirms
successful testing of the product by affixing to it the mark.
Endress+Hauser
7
Installation
Turbimax CUE21 / CUE22
4
Installation
4.1
Installation requirements
4.1.1
Measuring system
A complete measuring system comprises:
• Turbimax turbidimeter with installed cuvette and desiccant pouch
• Connecting tube with
– fittings for flow-through assembly
– Shutoff clamp
– Flow controller
– Vent screw for drain tube (in pressurized systems)
• Sensor interconnect cable
3
4
1
5
MODE
EXIT
E
6
Turbimax CUE21
2
7
8
A0025595
1
1
Shutoff clamp
2
Connection for intake tube
3
4
1)
8
Turbimax CUE21 (example)
5
Sensor interconnect cable
6
Vent for drain tube
Flow assembly
7
Connection for drain tube
Flow controller
8
Terminal box
1)
1)
OD 8 mm (0.31 inch), ID 4.75 mm (3/16 inch)
Endress+Hauser
Turbimax CUE21 / CUE22
Installation
Mounting dimensions
131.6 (5.16)
200 (8)
4.1.2
M6 (¼)
E
50.8 (2.0)
90.2 (35)
OUT
OUT
IN
347.2 (13.7)
207.7
(8.18)
M4
(3/16)
87.3 (3.44)
61.8
(2.43)
30.2
(1.19)
196.8 (7.75)
147.7 (5.81)
A0025606
2
Mounting dimensions - front view
A0025608
3
Mounting dimensions - lateral view
Dimensions in mm (inch)
4.2
Mounting the measuring device
4.2.1
Wall mounting
The device is designed for wall mounting. If wall mounting is not practicable, the device
can be mounted on any suitable level surface.
• Ensure that the temperature does not exceed the maximum permitted operating
temperature range (0 to 50 °C (32 to 122 °F)).
• Leave approx. 20 cm (8 inch) of free space above the device to ensure that there is
sufficient room for calibration and cuvette maintenance.(→ 2)
• Mount the device as close as possible to the sampling point to ensure fast response times
(within 2 to 3 m (6 to 10 ft)).
1.
Drill mounting holes for the M4 mounting screws to install the field terminal box.
2.
Drill mounting holes for the M6 mounting screws to install the device housing on top
of the terminal box.
3.
Mount the terminal box.
4.
Mount the rest of the device on top of the terminal box.
4.2.2
Installing the desiccant pouch
The device is equipped with a continuous vapor purge system. A replaceable desiccant
pouch in the lower portion of the instrument dries the air. System heat is used to warm the
Endress+Hauser
9
Installation
Turbimax CUE21 / CUE22
air. A fan located inside the device ensures that heated air circulates continuously around
the optical path and the flow-through cuvette.
1
3
2
1
Securing screws
2
Electronic component
3
Humidity indicator
4
Desiccant pouch
5
Base of housing
4
5
1
A0026679
4
Installing the desiccant pouch
1.
Unscrew the four corner screws (item 1).
2.
Remove the electronic component of the device (item 2).
3.
Open the bag protecting the desiccant pouch (item 4), and place the pouch with the
humidity indicator (item 3) in the base of the housing (item 5).
Ensure that you install the desiccant pouch immediately after opening the bag to
prevent premature degradation of the desiccant.
4.
Return the electronic component of the device to its original position.
5.
Tighten the four corner screws.
To prevent premature saturation of the desiccant, it is recommended that you keep
the measurement chamber covered at all times.
4.2.3
Plumbing the measuring device
• The device requires only a very low overpressure in the flow assembly to operate,
approx. 0.069 bar (1 psi).
• The flow controller is rated for a maximum overpressure of 13.8 bar (200 psi).
• The minimum flow rate for the flow-through cuvette is 100 ml/min to 1 l/min
(0.026 to 0.26 US gal/min).
• The maximum permitted medium temperature is 50 °C (122 °F).
10
Endress+Hauser
Turbimax CUE21 / CUE22
Installation
A0025699
5
Recommended plumbing
1
Connection for intake tube
7
Connection for drain tube
2
Shutoff clamp
8
Vent for drain tube
3
Connection for intake tube
9
Connection for drain tube
4
Connection for intake tube
10
Use
5
Connection for drain tube
11
Nut
6
Flow controller
12
Tubing (not supplied)
The device is equipped for plumbing using tubing with OD 8 mm (0.31 inch),
ID 4.75 mm (3/16 inch). If the device is going to be exposed to sunlight, opaque
tubing should be used to prevent algae growth.
Plumbing the device
Endress+Hauser
1.
Push the shutoff clamp (item 2) over one of the supplied tubes.
2.
Connect the tube to the upper part of the intake tubing connection (item 3) and to
the connection on top of the device (item 1).
3.
Push the flow controller (item 6) over the second tube supplied.
4.
Connect the tube to the upper part of the drain tubing connection (item 7) and to the
connection on top of the device (item 5).
5.
Connect a tube with an internal diameter of 4.75 mm (3/16 inch) to the lower part
of the intake tubing connection (item 4) to supply the sensor with a reliable sample
flow.
6.
Connect a tube with an internal diameter of 4.75 mm (3/16 inch) to the lower part
of the drain tubing connection (item 9) to route the sensor drain to a suitable drain
on site.
The vent in the drain tube (item 8) allows for pressure compensation, thus
helping to prevent bubble formation in the cuvette.
7.
If your CUE21 / CUE22 is integrated in a high-pressure system,
insert the 6:32 seal screw provided into the vent hole and tighten it.
11
Installation
Turbimax CUE21 / CUE22
4.3
12
Post-installation check
1.
After mounting, check all the connections to ensure they are secure and leak-tight.
2.
Check all cables and hoses for damage.
3.
Check whether the cables are routed such that they are free from electromagnetic
interference influences.
Endress+Hauser
Turbimax CUE21 / CUE22
Electrical connection
5
Electrical connection
LWARNING
Device is live
Incorrect connection may result in injury or death.
‣ The electrical connection may be performed only by an electrical technician.
‣ The electrical technician must have read and understood these Operating Instructions
and must follow the instructions contained therein.
Prior
to commencing connection work, ensure that no voltage is present on any cable.
‣
5.1
Connection requirements
‣ The device is equipped with a power supply of 100 to 240 V AC, 47 to 63 Hz. Before
connecting the device, verify that the line voltage matches these specifications.
‣ Place a circuit breaker upstream from the mains connection to facilitate maintenance
work.
Q2
R2
D2
D1
R3
1
EMCF1
Q1
R4
R1
RLY2
RLY1
4
TB2
TB1
TB3
1
5
NO
NC
ALARM
COM
2
NO
NC
100-240
VAC
90-250 VAC
COM
ALARM
47-63 Hz
+ SHLD
A
B
4-20mA/
RS-485
L1
L6
L2
WHT
RED
L5
BRN
ORG
TB4
L3
BLK
GRN
L7
2
YEL
SHLD
ISL1
L4
6
7
3
8
A0025701
Endress+Hauser
6
Field terminal box
1
Terminal strip, power supply
5
2
Power cable strain relief
6
Sensor wiring
3
Power cable gland
7
4 to 20 mA / RS485 cable gland
4
Alarm terminal strip, 240 VA, 2 A
8
Alarm cable gland
4 to 20 mA / RS485 terminal strip
13
Electrical connection
Turbimax CUE21 / CUE22
5.2
Connecting RS-485 cables
The digital RS-485 half-duplex (2-wire) interface operates with differential levels that are
not susceptible to electrical interference. That is why cable lengths of up to 914 m (3000
ft) can be used. The last device on each bus may require terminating with a 120-ohm
resistor to eliminate signal reflection on the line.
‣ Do not run the RS-485 cables in the same conduit as the power cable.
If you are using the device in a process control system without galvanically isolated
inputs, you must use a passive barrier e.g. RB223 (see "Accessories") to connect the
CUE21/22.
5.3
Connecting relays
The relays for alarm 1 and alarm 2 are mechanical relays rated at 240 V AC and 2 A.
‣ Please note that the relays are labeled NO (Normally Open), NC (Normally Closed) and
C (Common).
The alarm relays are configured to be "fail-safe" i.e. under normal operating conditions, the
device is supplied with current and is not in an alarm condition. How these alarm relays
function is described in the "Configuring the alarms" section.
5.4
Connecting the current output
The 4-20 mA output is supplied by a 15 V DC power source and can operate recording
devices with loads of up to 600 Ω. This output is isolated from line power and earth
ground.
‣ Do not run the analog current cables in the same conduit as the power cable.
How this output works is described in the "Selecting the output" section.
5.5
Ensuring the degree of protection
Only the mechanical and electrical connections which are described in these instructions
and which are necessary for the required, designated use, may be carried out on the device
delivered.
‣ Exercise care when carrying out the work.
Otherwise, the individual types of protection (Ingress Protection (IP), electrical safety, EMC
interference immunity) agreed for this product can no longer be guaranteed due, for
example, to covers being left off or cable (ends) which are loose or insufficiently secured.
5.6
Post-connection check
Device condition and specifications
Notes
Is there any external damage to the device and cables?
Visual inspection
Electrical connection
Notes
Are the installed cables strain-relieved and not twisted?
Visual inspection
Cable run without loops and cross-overs?
Is the cable type run completely isolated at the point of installation?
Power cables / signal lines
Are all the screws terminals properly tightened?
14
Endress+Hauser
Turbimax CUE21 / CUE22
Endress+Hauser
Electrical connection
Electrical connection
Notes
Are all cable entries mounted, tightened and leak-tight?
For lateral cable entries, make sure the
cables loop downwards to allow water
to drip off
Are the PE distributor blocks grounded (if present)?
Grounding is carried out at the point of
installation.
15
Operation options
Turbimax CUE21 / CUE22
6
Operation options
6.1
Overview
6.1.1
Display and operating elements
1
AUTO
Å
2
CAL
00000
..:.
CONFIG
3
0000 OFFSET
MODE
4
E
EXIT
5
6
7
8
A0025727
1
Displays turbidity value and user guidance
during configuration
5
Icon indicates use of an access code
2
Indicates operating mode: AUTO
(measurement), CAL (calibration), CONFIG
(configuration)
6
Indicates OFFSET mode
3
Displays error message and user guidance
7
Confirms a selected option or mode
4
Switches between modes
8
Keys used to change settings
6.1.2
Key functions
Switches between the operating modes of the device: CAL, CONFIG and AUTO (measurement)
Confirms the highlighted or selected option or mode
Scrolls up through the menu.
Increases numerical values.
Scrolls down through the menu.
Decreases numerical values.
6.2
Operating concept
6.2.1
Access code
The device is equipped with a security feature. An access code protects it from
unintentional or undesired changes to the configuration and to the calibration data. The
access code can be enabled in the configuration mode.
16
Endress+Hauser
Turbimax CUE21 / CUE22
Operation options
A0026267
7
Access protection enabled: Prompt to enter code
‣ If access protection is enabled:
Enter the code 333.
You can access the CAL or CONFIG menus.
The access code cannot be changed.
6.2.2
Menu structure
The device has three operating modes that can be selected using the
• AUTO
Standard operating mode for displaying current measured values
• CAL
Calibration mode for performing calibration procedures
• CONFIG
Setup mode for configuring customer-specific settings
key:
Configuration menu (CONFIG)
The configuration menu is split into several submenus. The following submenus are
available:
• Output selection
• RS-485 port configuration
• Alarm configuration
• Security access configuration
• Offset configuration
• Advanced settings
The extended settings are grouped together to prevent them from being modified
accidentally:
– Response speed
– Displayed resolution
– LCD backlight brightness
– Displayed units
– Ultrasonic cleaning
– RS485 parameters
– Desiccant alarm
For a detailed explanation of the submenus, see the "Instrument configuration"
section. (→ 19)
Endress+Hauser
17
Commissioning
Turbimax CUE21 / CUE22
7
Commissioning
7.1
Function check
‣ Verify that all connections are correct.
‣ In particular, verify that all tubes are firmly attached and are not leaking.
‣ Ensure that the supply voltage is within the permitted range of 100 to 240 V AC!
7.2
Switching on the measuring device
Familiarize yourself with the operation of the device prior to switching it on. To do so, read
the "Basic safety instructions" and "Operation" sections in particular.
18
1.
Switch on the power supply to the device.
2.
Wait until the device has warmed up (typically 45 to 60 minutes when
commissioning the device for the first time).
When a continuous process stream is flowing through the device, the device will
output the turbidity level measured in the sample to the LC display. In addition,
the equivalent signal is output on the analog (4-20 mA) or digital output,
depending on the options selected.
3.
Configure the device according to your needs.
Endress+Hauser
Turbimax CUE21 / CUE22
Operation
8
Operation
8.1
Configuring the measuring device
8.1.1
Selecting the output
Function
Options
Output (O/P)
• 4-20
• 485
• OFF
AUTO
4-20
CAL
Å
CONFIG
Info
O/P
4-20
4 mA value
0 to 1000 NTU
Set the lower turbidity limit value (LOLM) that
corresponds to the 4 mA output level.
20 mA value
0 to 1000 NTU
Set the upper turbidity value (UPLM) that
corresponds to the 20 mA output level.
Baud
•
•
•
•
•
Set the correct baud rate for operation of the I/O
port.
Addr
1 ... 255
Select the desired device address.
MBUS
• ASCII
• RTU
Select ASCII or RTU.
485
8.1.2
Configuring the alarms
Function
Options
Alarm 1 (ALM1)
• HI
• LO
• OFF
AUTO
CAL
Å
CONFIG
HI
Info
ALM1
Set point (S/P)
0 to 1000 NTU
Set the desired alarm set point.
Delay on (DLY▴)
1 to 30 s
Set the desired number of seconds for the "Delay
on" function.
Delay off (DLY▾)
1 to 30 s
Set the desired number of seconds for the "Delay
off" function.
Alarm 2 (ALM2)
Set point (S/P)
Endress+Hauser
1200
2400
4800
9600
19200
• HI
• LO
• OFF
0 to 1000 NTU
Set the desired alarm set point.
19
Operation
Turbimax CUE21 / CUE22
Function
Options
Info
Delay on (DLY▴)
1 to 30 s
Set the desired number of seconds for the "Delay
on" function.
Delay off (DLY▾)
1 to 30 s
Set the desired number of seconds for the "Delay
off" function.
Function
Options
Info
Alarm 1 (ALM1)
• On
• OFF
If the security feature is switched on to restrict
access, you are required to enter an access code if
you wish to switch from AUTO to a different
operating mode. The only valid code is 333. This
code cannot be changed.
8.1.3
Access code
AUTO
On
CAL
Å
CONFIG
8.1.4
CODE
Advanced settings
Function
Options
Info
Extended settings (EXTD)
• On
• OFF
Switch on the "Extended settings" function to access
the configuration of the following options:
• Response speed
• Displayed resolution
• LCD backlight brightness
• Units
• Ultrasonic cleaning
• RS-485 parameters
• Desiccant alarm
Function
Options
Info
Response speed (RESP)
1 to 100 %
Factory setting: 10
Select the desired response speed for the NTU
values that are displayed and output.
Select the lowest speed (highest number) to avoid
displaying the effects of air and other anomalies.
Select the highest speed if there are rapid changes
taking place in the process that require monitoring.
The displayed number is a relative speed. The
approximate response time (in seconds) is
calculated by multiplying the displayed number by
5.
Function
Options
Info
Resolution (RES)
1 to 0.0001
For readings below 10 NTU the device can display
Factory setting: 0.01 values with a resolution of up to four digits to the
right of the decimal point.
AUTO
On
CAL
Å
CONFIG
EXTD
Response speed
AUTO
10
CAL
Å
CONFIG
RESP
Displayed resolution
AUTO
CAL
Å
20
CONFIG
0.01
RES
Endress+Hauser
Turbimax CUE21 / CUE22
Operation
LCD backlight brightness
Function
Options
Info
LCD backlight brightness (BRT)
1 to 10
Factory setting: 8
Adjust the backlighting for the display if necessary.
Function
Options
Info
Units (UNIT)
• NTU
• FNU
AUTO
8
CAL
Å
CONFIG
BRT
Units
AUTO
CAL
Å
CONFIG
NTU
Select the desired unit for displaying the turbidity
level: NTU (Nephelometric Turbidity Units) or FNU
Factory setting: NTU (Formazin Nephelometric Units).
UNIT
Ultrasonic cleaning
The "Ultrasonic cleaning" option is used for continuous cleaning of the flow-through
cuvette. It is not intended to clean cuvettes that are already dirty, or to replace manual
cleaning entirely. However, this system will increase the cleaning intervals dramatically.
To operate correctly, the use of a cuvette with an ultrasonic transducer is required. The
system transmits an ultrasonic frequency through a spring to a piezo transducer that is
joined to the base of the flow-through cuvette. The cuvette's detection system works only
in AUTO mode.
The system can detect the following error conditions:
• Incorrect cuvette installed
• Error has occurred in the transducer
• Transducer not making contact with the spring
If an error has occurred, the message "CLN" appears in the lower part of the display. The
alarms are set and 2mA is sent to the current output. The ultrasonic cleaning function is
switched on as standard. After a cuvette has been inserted, the message "Dry" appears for a
duration of 30 minutes in the lower part of the display. During this time, the desiccant
removes residual moisture from the ultrasonic transducer and no ultrasonic cleaning takes
place. At the same time, the error message "CLN" is reset and appears only after the 30
minutes have elapsed.
Function
Options
Info
Ultrasonic cleaning (CLN)
• On
• OFF
The ultrasonic cleaning function can be turned off
if desired.
AUTO
CAL
Å
Endress+Hauser
CONFIG
CLN
On
On
21
Operation
Turbimax CUE21 / CUE22
RS-485 parameters
Using basic programs, such as the Hilgraeve HyperTerminal (included in most Microsoft
Windows software packages), the Turbimax can provide basic communication functions.
You could also use Visual Basic or other programs. The standard communication
parameters are: 8 bits, no parity and 1 stop bit. You can modify these parameters using the
following menus. They will appear only if RS-485 was selected as the output (→ 19).
Function
Options
Info
Bits (BITS)
• 7
• 8
Select the required number of data bits for your
communication software.
AUTO
CAL
Å
CONFIG
BITS
8
Parity (PRTY)
Factory setting: 8
• nOnE
• ODD
• E
Select the required parity bit for your
communication software.
Factory setting:
nOnE
Stop
• 1
• 2
Select the required stop bit for your communication
software.
Factory setting: 1
Sample communication
The master computer requests a report from the Turbimax.
• To this end, the master transmits the following to the Turbimax:
– Byte #1 the initial character ":" in ASCII or "3A" Hex
– Byte #2 the address of the Turbimax being queried
– Byte #3, 4 "CR LF" in ASCII or "0D 0A" in Hex
• The Turbimax responds with:
– The same initial character ":" in ASCII or "3A" Hex
– Its address
– The display value for the turbidity content
– The unit
Communication would proceed as follows:
‣ The master computer requests a report from address 1: : 1 CRLF
The Turbimax, which is located at address 1, responds with: : 001 0.0249 NTU
Desiccant alarm
If the humidity detector in the Turbimax indicates that the humidity of the internal
environment may lead to condensation, the device outputs the warning DESC. If desired,
the desiccant warning can activate the alarms and send 2 mA to the current output.
Function
Options
Info
Desiccant alarm (DESC)
• OFF
• On
Select "On" to activate the alarms as soon as the
desiccant can no longer fulfil its function.
AUTO
CAL
Å
22
CONFIG
OFF
Factory setting: OFF
DESC
Endress+Hauser
Turbimax CUE21 / CUE22
Operation
8.2
Instrument calibration
The device was calibrated and tested prior to leaving the factory. Therefore, it is possible to
use it directly out of the box. Under normal conditions, recalibration is recommended at
least once every three months. 1)
Types of calibration
• Complete three-point calibration
This calibration is used if accuracy is required across the entire measuring range of the
device, from 0.02 to 1000 NTU (CUE21) or 0.02 to 100 NTU (CUE22). If accuracy is
required only in the range below 10.0 NTU, e.g. for drinking water, calibration may also
be performed using only a 10.0 NTU and 0.02 NTU calibration standard.
• Offset
This method of calibration can be used if accuracy is required only in the immediate
vicinity of the turbidity value of the sample.
• The relay contacts remain in the last valid condition while the device is in
calibration mode.
• The device has a timeout function which causes it to revert automatically to AUTO
mode after 15 minutes of inactivity.
• During calibration, the fan located inside the device is switched off in order to
extend the service life of the desiccant. The fan is switched on during calibration
countdowns and after the device returns to AUTO mode or after five minutes
(whichever comes first). It is recommended that the measurement chamber be kept
covered during calibration and that the flow-through cuvette be inserted
immediately after calibration to prevent premature saturation of the desiccant.
8.2.1
Calibrations standards
To achieve the specified level of accuracy across the entire measuring range of the device,
we recommend the following calibration standards:
• 0.02 NTU
• 10.0 NTU
• 1000 NTU (CUE21)
100 NTU (CUE22)
These calibrations standards are more stable than formazin and have a shelf life of 12
months. Prior to calibration, verify the expiry dates to ensure that these calibration
standards have not expired. If you are using formazin to calibrate the device, ensure that a
fresh formazin stock solution is used so that the specified accuracy is achieved.
A a shelf life of 12 months is guaranteed for the 10.0 NTU calibration standard only if
it is stored in the plastic bottle. Once it is transferred to a glass cuvette, it must be
used for calibration immediately and then disposed of.
8.2.2
Indexing cuvettes
To achieve the highest possible level of accuracy and to allow for normal scratches and
signs of wear and tear on the cuvette glass when calibrating, we recommend that the
cuvettes be indexed. Calibration standards and standard kits purchased from Endress
+Hauser are supplied with indexing rings
The following steps allow repeatable indexing of calibration standards:
1)
1.
Shut off the flow using the shutoff clamp provided.
2.
Remove the flow assembly with the flow-through cuvette.
3.
Insert the calibration standard while the device is in AUTO mode.
The EPA recommends that online turbidimeters be calibrated using a calibration standard at least once every three months if they are used for
EPA reporting.
Endress+Hauser
23
Operation
Turbimax CUE21 / CUE22
4.
Slowly rotate the calibration standard inside the optical path one complete revolution
(360°).
While rotating the standard slowly, observe the measured value. Locate the
position at which the lowest reading is displayed.
5.
With the calibration standard positioned at the point with the lowest reading, install
the indexing ring over the cap on the calibration standard . Make sure that the
pointer of the indexing ring faces directly forward.
When using the standards in future, always insert them in such a way that the pointer of
the indexing ring faces forward. Slowly rotate the calibration standard back and forth by
approx. 5° to find the lowest value. The calibration standard is now indexed and ready for
use.
8.2.3
Three-point calibration
First calibration point
1.
Press the MODE/EXIT key until the calibration function is displayed.
The arrow appears beside "CAL". The lower part of the display alternates between
1000 (CUE21) or 100 (CUE22) (the value of the requested calibration standard)
and
. The upper part of the display shows the current measured value.
If you wish to perform a two-point calibration for drinking water, press the
key to bypass the 1000 NTU or 100 NTU and to proceed with the second
calibration point.
2.
If the flow assembly with the flow-through cuvette has not yet been removed,
remove it now.
3.
Insert the requested calibration standard (1000 NTU or 100 NTU) so that the pointer
of the indexing ring faces directly forward.
4.
Find the lowest measured value by rotating the standard slowly back and forth by 5°.
Leave the standard in the position with the lowest reading.
5.
Press the key to accept the calibration.
In the lower part of the display, the stabilization time for this calibration step is
counted down.
Second calibration point
24
1.
The lower part of the display alternates between 10 and
, thereby requesting the
10.0 NTU calibration standard.
, press the or key until the
If the display does not alternate between 10 and
specified information appears.
2.
Insert the requested calibration standard so that the pointer of the indexing ring
faces directly forward.
Endress+Hauser
Turbimax CUE21 / CUE22
Operation
3.
Find the lowest measured value by rotating the standard slowly back and forth by 5°.
Leave the standard in the position with the lowest reading.
4.
Press the key to accept the calibration.
In the lower part of the display, the stabilization time for this calibration step is
counted down.
Third calibration point
The lower part of the display alternates between 02 and
0.02 NTU calibration standard.
, thereby requesting the
1.
Insert the requested calibration standard so that the pointer of the indexing ring
faces directly forward.
2.
Find the lowest measured value by rotating the standard slowly back and forth by 5°.
Leave the standard in the position with the lowest reading.
3.
Press the key to accept the calibration.
In the lower part of the display, the stabilization time for this calibration step is
counted down.
When calibration is complete, the device returns to AUTO mode.
8.2.4
Offset
In certain instances, it can be useful to use an offset instead of a three-point calibration.
However, an offset cannot be used in lieu of regular calibration. Using this method, the
device will provide accurate turbidity values only in the immediate vicinity of the sample
value and not for the entire measuring range.
The OFFSET icon is illuminated as soon as an offset is used. The maximum offset is
1.00 NTU. If the device deviation is greater than 1 NTU, a full calibration is
recommended.
Configuring the offset
Endress+Hauser
1.
Take a sample of the process medium that is being monitored by the device, and
record the turbidity values reported by the device.
2.
Measure the turbidity of the sample using a laboratory turbidimeter (e.g. Turbimax
CUE23 / CUE24).
3.
Compare the turbidity value reported by the device to the value measured in the lab.
If the values are very close, no offset adjustment is required and the procedure
can be stopped at this point.
If the values differ substantially (but by less than 1 NTU), continue the procedure
to configure an offset.
4.
Select the offset function by pressing the MODE/EXIT key until the arrow appears
next to CONFIG.
25
Operation
Turbimax CUE21 / CUE22
5.
Press the key until OFST is displayed in the lower line.
At this point, the upper line of the display indicates the operating status of the
offset function. If the function is switched off, switch it on by pressing or .
6.
The offset is the difference between the NTU value measured by the device and the
value measured by the lab device. Example: CUE21 / CUE22 measures 0.016 NTU,
while the lab device calculates a value of 0.012 NTU for the same sample. Entering
an offset of -0.04 results in the CUE21 / CUE22 also displaying 0.012 NTU.
Select the desired offset value using the or keys. Press the key to accept the
value.
This completes offset configuration. The device is still in configuration mode (CONFIG).
Press the MODE/EXIT key to return to AUTO mode.
26
Endress+Hauser
Turbimax CUE21 / CUE22
Diagnostics and troubleshooting
9
Diagnostics and troubleshooting
9.1
General troubleshooting
9.1.1
Troubleshooting instructions
The device performs continuous diagnostic monitoring. Any errors are displayed in a
queue in the bottom line of the display.
The Turbimax provides three levels of fault detection:
• Warning
A warning is simply a message on the display indicating that a problem exists. No alarms
are activated. If, for example, the desiccant alarm function is disabled and the desiccant
becomes saturated, the warning DESC will appear on the display.
• Error
An error indicates a fault or a problem that can usually be corrected by the operator. This
includes, for example, a lamp outage (LAMP) or an incorrect calibration (CAL). If an
error occurs, both alarm relays are activated and the current output is maintained at 2
mA. While the device continues to display values, you should not rely on these values as
their accuracy is not known.
• Failure
A failure is a system fault. This is a problem that cannot be corrected by the operator.
The unit must be returned to the factory for service. These failures consist of failures in
the CPU, A/D, EEPROM or other devices internal to the instrument. If a failure occurs,
the device will not function properly and will display the message FAIL in the bottom
line of the display. Both alarm relays are activated and the current output is maintained
at 2 mA.
9.1.2
Process-related errors
Error
Possible cause
Tests or remedial action
Display values are higher than
expected
Bubbles in solution
Ensure that vent in drain tube is open and
not obstructed (→ 10)
Condensate or leaky cuvette
Check flow-through cuvette for
condensate or leaks
Flow-through cuvette dirty
Clean cuvette (→ 29).
Device is no longer correctly
calibrated
Recalibrate device (→ 23)
Bubbles in solution
see above
Debris in flow-through cuvette
Remove debris from cuvette
Display values are lower than
expected
Device is no longer correctly
calibrated
Recalibrate device (→ 23)
Upper part of display flashing
Sample has exceeded permitted
range
Check sample. Turbidity of sample may be
too high to be read by the device.
Display values are incorrect
9.2
Endress+Hauser
Diagnostic information on local display
Diagnostic
message
Possible cause
Tests or remedial action
MA
4 to 20 mA loop is open
Check wiring (→ 13, → 19).
DESC
Desiccant pouch saturated
Replace desiccant pouch (→ 30).
LAMP
Lamp failed
Have lamp replaced. Contact sales office.
27
Diagnostics and troubleshooting
28
Turbimax CUE21 / CUE22
Diagnostic
message
Possible cause
Tests or remedial action
FLOW
Sample flow has stopped
Restore flow
CLN
Ultrasonic transducer in flowthrough cuvette not making contact
Rotate the flow-through cuvette slightly to improve
the connection. If the problem persists, replace the
cuvette.
Flow-through cuvette removed
Insert flow-through cuvette
FAIL
Major system fault
Contact sales office.
DRY
Drying period
Disappears automatically after a period of 30
minutes.
Endress+Hauser
Turbimax CUE21 / CUE22
Maintenance
10
Maintenance
Take all the necessary precautions in time to ensure the operational safety and reliability
of the entire measuring system.
NOTICE
Effects on process and process control
‣ When carrying out any work on the system, take into account possible repercussions
for process control or the process itself.
‣ For your own safety, only use genuine accessories. With genuine parts, the function,
accuracy and reliability are also ensured after maintenance work.
10.1
Cleaning the flow-through cuvette
The cuvettes should always be clean and free of marks and scratches.
1.
Clean the inside of the cuvette using a cleaning solution.
2.
Rinse the cuvette several times using distilled or deionized water.
3.
If external dirt or scratches cannot be removed:
Replace the cuvette. To do so, read the next section.
10.2
Replacing the flow-through cuvette
NOTICE
Moisture on the cuvette or transducer
High risk of damage to sensor electronics or transducer
‣ Check the cuvette prior to installation. It must be completely dry.
1
2
MODE
EXIT
E
A0026688
8
1
2
Endress+Hauser
Replacing the cuvette
Flow assembly
Cuvette
1.
Shut off flow at the Turbimax using the shutoff clamp provided.
2.
Remove the flow assembly (item 1) from the device.
3.
Unscrew the old cuvette (item 2).
4.
Verify that the new cuvette is clean, dry and free of scratches or other marks.
Carefully screw the cuvette into the flow assembly making sure not to leave
fingerprints on the cuvette.
5.
Insert the cuvette with the flow assembly into the device and lock the flow assembly.
After inserting the clean cuvette, the message "DRY" is displayed during the
drying phase.
29
Maintenance
Turbimax CUE21 / CUE22
10.3
Replacing the desiccant pouch
Correct use of the desiccant supplied is essential in maintaining the performance of the
device. In addition, the housing seal must be maintained to guarantee the standard service
life of the desiccant. Inspect this seal each time the desiccant pouch is replaced.
Replace the desiccant when the device outputs a desiccant alarm (DESC).
1
1
Humidity indicator
2
Desiccant pouch
3
Housing seal
4
Base of housing
2
3
4
A0026690
9
Replacing desiccant
1.
Switch off the power supply to the device.
2.
Unscrew the four corner screws (item 1).
3.
Remove the device's electronic component.
4.
Remove the used desiccant pouch (item 2).
5.
Check the housing seal (item 3) on the base of the housing.
Replace it if necessary.
6.
Open the bag protecting the new desiccant pouch and place the desiccant pouch and
the humidity indicator (item 1) in the base of the housing (item 4).
Ensure that you install the desiccant pouch immediately after opening the bag to
prevent premature degradation of the desiccant.
7.
Return the electronic component of the device to its original position.
8.
Tighten the four corner screws.
9.
Switch the power supply to the device back on.
The device starts up. Wait for initialization.
10. To speed up detection of the new desiccant, the device must be reset.
This is done by removing the sensor interconnect cable from the device for 2 seconds
and then reconnecting it.
30
Endress+Hauser
Turbimax CUE21 / CUE22
Repairs
11
Repairs
11.1
Spare parts
Description and contents
Order number
Electronic service module for CUE21, infrared
71030103
Electronic service module for CUE22, white light
71030104
Terminal box CUE21 / CUE22
71030105
Desiccant with indicator card, refill pack
51518578
Tubing kit including:
• 1 shutoff clamp
• 1 flow controller
• 2 connecting tubes with fittings for flow assembly
• Vent for drain tube
51518579
Replacement cuvette with ultrasonic transducer for use of the ultrasonic cleaning function
51518576
Pressure regulator for CUE21/22
71085512
11.2
Replacing the electronics module
1
2
1
Securing screws
2
Electronics module
3
Base of housing
3
1
A0026681
10
Replacing the electronics module
1.
Unscrew the four corner screws (item 1).
2.
Remove the electronics module (item 2) from the housing base (item 3).
3.
Place the new electronics module on the housing base.
4.
Secure using the four screws.
11.3
Replacing the lamp
The lamps in the device have an operating life of approx. 10 years.
Should the lamp need to be replaced, please contact your sales office.
Endress+Hauser
31
Repairs
Turbimax CUE21 / CUE22
11.4
Return
The product must be returned if repairs or a factory calibration are required, or if the
wrong product was ordered or delivered. As an ISO-certified company and also due to legal
regulations, Endress+Hauser is obliged to follow certain procedures when handling any
returned products that have been in contact with medium.
To ensure swift, safe and professional device returns, please read the return procedures
and conditions at www.endress.com/support/return-material.
11.5
Disposal
The device contains electronic components and must therefore be disposed of in
accordance with regulations on the disposal of electronic waste.
Observe the local regulations.
32
Endress+Hauser
Turbimax CUE21 / CUE22
Accessories
12
Accessories
The following are the most important accessories available at the time this
documentation was issued. For accessories not listed here, please contact your service
or sales office.
12.1
Calibration solutions
Calibration kit CUE21 / CUE23 / CUE24
• Complete measuring range:
– 0.02 NTU
– 10.0 NTU
– 1000 NTU
• Order No.: 51518580
Calibration kit CUE22
• Complete measuring range:
– 0.02 NTU
– 10.0 NTU
– 100 NTU
• Order No.: 71030102
12.2
Flow chamber
Flow chamber CUE21 / CUE22
• To suppress air bubbles
• Order no.: 51518575
12.3
Passive barrier
RB223
• Loop-powered passive barrier
• For galvanic isolation of active signal circuits (0 to 20 mA)
• Product Configurator on the product page: www.endress.com/rb223
Technical Information RB223, TI00132R
Endress+Hauser
33
Technical data
Turbimax CUE21 / CUE22
13
Technical data
13.1
Input
Measured variables
Turbidity
Measuring ranges
CUE21:
CUE22:
13.2
0 to 1000 NTU / FNU
0 to 100 NTU / FNU
Output
Output signal
0/4 to 20 mA
Signal on alarm
Error current: 2 mA
Load
max. 600 Ω
Relay output
Switching voltage:
Switching current:
Communication interface
Bi-directional RS-485, optional Modbus
Limit value and alarm
functions
Setpoint adjustment:
CUE21
CUE22
Alarm delay:
13.3
max. 240 VAC
max. 2 A
0 to 1000 NTU
0 to 100 NTU
0 to 30 s
Power supply
Supply voltage
100 to 240 VAC, 47 to 63 Hz, 80 VA switching power supply
Overvoltage protection
Category II
Protection class
2
13.4
Performance characteristics
Response time
adjustable from 1 to 100 % (approx. 5 to 500 s)
Reference temperature
25 °C (77 °F)
34
Endress+Hauser
Turbimax CUE21 / CUE22
Technical data
Measured value resolution
0.0001 NTU (below 10 NTU)
Maximum measured error
below 40 NTU:
above 40 NTU:
Repeatability
±2 % of display value or ±0.02 NTU - depending on
which value is higher
±5 % of display value
±1% of reading
13.5
Environment
Ambient temperature
2 to 50 °C (36 to 120 °F)
Storage temperature
-20 to +60 °C (0 to 140 °F)
Humidity
0 to 95 %, non-condensating
Altitude
<2000 m (6500 ft)
Degree of protection
IP66
Pollution degree
The product is suitable for pollution degree 2.
13.6
Process
Process temperature
2 to 50 °C (36 to 120 °F)
Process pressure
max. 13.78 bar (200 psi), controlled by integrated flow controller
Flow
0.1 to 1 l/min (0.026 to 0.26 US gal/min)
13.7
Mechanical construction
Dimensions
H x W x D:
Weight
2.0 kg (4.4 lbs)
Endress+Hauser
347 x 208 x 197 mm (13.7 x 8.18 x 7.75 inch)
35
Technical data
Materials
Light sources
36
Turbimax CUE21 / CUE22
Housing:
ABS
Flow assembly:
Nylon
Cuvette:
Borosilicate glass
Cuvette seal:
Silicone
Fittings (assembly):
Polypropylene
Locking pins (assembly):
Stainless steel (AISI 304 or AISI 303)
Intake tubing:
Stainless steel (AISI 316)
CUE21:
CUE22:
Infrared LED, 860 nm
White light tungsten lamp, ~600 nm, 2250 K
Endress+Hauser
Turbimax CUE21 / CUE22
Index
Index
A
Access code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Advanced settings . . . . . . . . . . . . . . . . . . . . . . . . . . .
Altitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Ambient temperature . . . . . . . . . . . . . . . . . . . . . . . .
16
20
35
35
N
C
Certificates and approvals . . . . . . . . . . . . . . . . . . . . . . 7
Check
Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Terminal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Cleaning the flow-through cuvette . . . . . . . . . . . . . . . 29
Connection requirements . . . . . . . . . . . . . . . . . . . . . . 13
D
Degree of protection . . . . . . . . . . . . . . . . . . . . . . 14, 35
Desiccant pouch
Installing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Replacing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Diagnosis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Diagnostic information . . . . . . . . . . . . . . . . . . . . . . . 27
Display and operating elements . . . . . . . . . . . . . . . . . 16
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . 13
F
Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
H
Humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
I
Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . . 6
Installation
Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Installation requirements . . . . . . . . . . . . . . . . . . . . 8
Mounting the measuring device . . . . . . . . . . . . . . . 9
Installation requirements
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . 8
Mounting location . . . . . . . . . . . . . . . . . . . . . . . . . 9
L
Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measured error . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measured value resolution . . . . . . . . . . . . . . . . . . . . .
Measured variables . . . . . . . . . . . . . . . . . . . . . . . . . .
Measuring device
Calibrating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Configuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Endress+Hauser
Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
O
Operating concept . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Operating keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Operation
Configuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Menu structure . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Operation options . . . . . . . . . . . . . . . . . . . . . . . . . 16
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Overvoltage protection . . . . . . . . . . . . . . . . . . . . . . . 34
P
Pollution degree . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Process pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Process temperature . . . . . . . . . . . . . . . . . . . . . . . . . 35
Process-related errors . . . . . . . . . . . . . . . . . . . . . . . . 27
Product identification . . . . . . . . . . . . . . . . . . . . . . . . . 6
R
E
M
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Operation options . . . . . . . . . . . . . . . . . . . . . . . . . 16
Switching on . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Measuring ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Menu structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
29
35
35
34
23
19
Reference temperature . . . . . . . . . . . . . . . . . . . . . . .
Repairs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Replacing
Desiccant pouch . . . . . . . . . . . . . . . . . . . . . . . . . .
Electronics module . . . . . . . . . . . . . . . . . . . . . . . .
Flow through cuvette . . . . . . . . . . . . . . . . . . . . . .
Light . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Replacing the electronics module . . . . . . . . . . . . . . . .
Replacing the flow-through cuvette . . . . . . . . . . . . . .
Response time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
34
31
35
30
31
29
31
31
29
34
32
S
Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . 35
Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
T
Technical data
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Mechanical construction . . . . . . . . . . . . . . . . . . . .
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance characteristics . . . . . . . . . . . . . . . . .
35
34
35
34
34
37
Index
Turbimax CUE21 / CUE22
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
U
Usage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
38
Endress+Hauser
*71292787*
71292787
www.addresses.endress.com