Endress + Hauser Operating Instructions STIP-scan CAM74/CAS74
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Operating Instructions
STIP-scan CAM74/CAS74
System for online measurement of nitrate, CODeq, BODeq,
TOCeq, SAC, total solids, sludge volume, sludge index and
turbidity
BA423C/07/en/05.09
71087097
Valid from:
Software version 19
Endress+Hauser
STIP-scan CAM74/CAS74
Table of contents
1
Safety instructions . . . . . . . . . . . . . . . . 4
6
Commissioning . . . . . . . . . . . . . . . . . . 59
1.1
1.2
1.3
1.4
1.5
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Installation, commissioning and operation . . . . . . . .
Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . .
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Notes on safety icons and symbols . . . . . . . . . . . . . .
4
4
4
5
5
6.1
6.2
6.3
Installation and functional check . . . . . . . . . . . . . . 59
Quick setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
7
Maintenance . . . . . . . . . . . . . . . . . . . . 64
2
Identification . . . . . . . . . . . . . . . . . . . . 6
7.1
2.1
2.2
2.3
Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1.1 Product structure for transmitter CAM74 . .
2.1.2 Product structure for sensor CAS74 . . . . . . .
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . .
Certificates and approvals . . . . . . . . . . . . . . . . . . . .
Maintenance of complete measuring point . . . . . . .
7.1.1 Cleaning the transmitter . . . . . . . . . . . . . .
7.1.2 Cleaning the sensor . . . . . . . . . . . . . . . . . .
7.1.3 Leakage test . . . . . . . . . . . . . . . . . . . . . . . .
8
Accessories . . . . . . . . . . . . . . . . . . . . . 66
3
Installation . . . . . . . . . . . . . . . . . . . . . . 9
9
Troubleshooting . . . . . . . . . . . . . . . . . 67
3.1
3.2
3.3
9.1
9.2
9.3
9.4
9.5
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10
Technical Data. . . . . . . . . . . . . . . . . . . 70
3.4
Incoming acceptance, transport, storage . . . . . . . . . . 9
Installation conditions . . . . . . . . . . . . . . . . . . . . . . 10
Installation instructions . . . . . . . . . . . . . . . . . . . . . 11
3.3.1 Installation with upright post . . . . . . . . . . . 11
3.3.2 Installing the sensor wall-bracket . . . . . . . . 13
3.3.3 Installing the sensor . . . . . . . . . . . . . . . . . . 14
3.3.4 Connecting the flow assembly . . . . . . . . . . 15
Post-installation check . . . . . . . . . . . . . . . . . . . . . . 15
4
Wiring . . . . . . . . . . . . . . . . . . . . . . . . 16
4.1
4.2
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance characteristics . . . . . . . . . . . . . . . . . .
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Mechanical construction . . . . . . . . . . . . . . . . . . . .
4.4
Electrical connection . . . . . . . . . . . . . . . . . . . . . . .
Connecting sensor and transmitter (stainless steel
version) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.2.1 Electrical connection of transmitter . . . . . .
4.2.2 Connecting the signal outputs . . . . . . . . . .
4.2.3 Connecting the relay outputs . . . . . . . . . . .
4.2.4 Connecting the panel PC . . . . . . . . . . . . . .
Connecting sensor and transmitter
(aluminum version) . . . . . . . . . . . . . . . . . . . . . . . .
4.3.1 Electrical connection of transmitter . . . . . .
4.3.2 Connecting the relay and signal outputs . . .
4.3.3 Connecting the PROFIBUS . . . . . . . . . . . .
4.3.4 Connecting of the power . . . . . . . . . . . . . .
Post-connection check . . . . . . . . . . . . . . . . . . . . . .
10.1
10.2
10.3
10.4
10.5
10.6
10.7
5
Operation . . . . . . . . . . . . . . . . . . . . . . 25
5.1
5.2
Human interface . . . . . . . . . . . . . . . . . . . . . . . . . .
Onsite operation . . . . . . . . . . . . . . . . . . . . . . . . . .
5.2.1 MEASUREMENT menu . . . . . . . . . . . . . . .
5.2.2 TRACES menu . . . . . . . . . . . . . . . . . . . . .
5.2.3 CALIBRATION menu . . . . . . . . . . . . . . . .
5.2.4 SETTINGS menu . . . . . . . . . . . . . . . . . . . .
5.2.5 TEST menu . . . . . . . . . . . . . . . . . . . . . . . .
5.2.6 LANGUAGE menu . . . . . . . . . . . . . . . . . .
4.3
Endress+Hauser
6
7
7
8
8
16
16
17
18
19
20
64
64
64
65
67
68
69
69
69
70
70
70
71
71
71
71
Index . . . . . . . . . . . . . . . . . . . . . . . . . . 72
20
21
23
23
23
24
25
26
26
29
38
43
54
58
3
Safety instructions
STIP-scan CAM74/CAS74
1
Safety instructions
1.1
Designated use
STIP-scan is a system for online measurement of nitrate, CODeq, BODeq, TOCeq, SAC, total solids,
sludge volume, sludge index and turbidity.
The measuring system is particularly suitable for use in the following areas:
• Continuous monitoring of wastewater for organic pollution and/or nitrate
• Measurement of sludge parameters
• Special measurements in the spectral range from 200 to 680 nm
Any other use than the one described here compromises the safety of persons and the entire
measuring system and is not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.
1.2
Installation, commissioning and operation
Please note the following items:
• Installation, commissioning, operation and maintenance of the measuring system must only be
carried out by trained technical personnel.
Trained personnel must be authorized for the specified activities by the system operator.
• Electrical connection must only be carried out by a certified electrician.
• Technical personnel must have read and understood these Operating Instructions and must
adhere to them.
• Before commissioning the entire measuring point, check all the connections. Ensure that
electrical cables and hose connections are not damaged.
• Do not operate damaged products and secure them against unintentional commissioning.
Mark the damaged product as being defective.
• Measuring point faults may only be rectified by authorized and specially trained personnel.
• If faults can not be rectified, the products must be taken out of service and secured against
unintentional commissioning.
• Repairs not described in these Operating Instructions may only be carried out at the
manufacturer’s or by the service organization.
1.3
Operational safety
The assembly has been designed and tested in accordance with the latest industry standards and left
the factory in perfect functioning order.
Relevant regulations and standards have been met.
As the user, you are responsible for complying with the following safety conditions:
• Installation instructions
• Local prevailing standards and regulations.
Immunity to interference
This instrument has been tested for electromagnetic compatibility in industrial use according to
applicable European standards.
Protection against interference as specified above is valid only for an instrument
connected according to the instructions in these Operating Instructions.
4
Endress+Hauser
STIP-scan CAM74/CAS74
Safety instructions
1.4
Return
If a repair is needed, please clean the sensor and return it to your sales office.
When returning the item, please use the original packaging.
Please enclose the completed "Declaration of hazardous material and de-contamination" (copy the
second last page of these Operating Instructions) in the packaging together with the dispatch
documents. No repairs can be carried out without a completed declaration!
1.5
#
"
!
Notes on safety icons and symbols
Warning!
This symbol alerts you to hazards that can cause serious damage to the instrument or to persons if
ignored.
Caution!
This symbol alerts you to possible faults which could arise from incorrect operation. They could
cause damage to the instrument if ignored.
Note!
This symbol indicates important items of information.
Electrical symbols
%
&
)
*
b
Direct Current (DC)
A terminal at which DC is applied or through which DC flows.
Alternating Current (AC)
A terminal at which (sine-form) AC is applied or through which AC flows.
Ground connecting
A terminal which, from the user’s point of view, is already grounded using a grounding system.
Protective ground terminal
A terminal which must be grounded before other connections may be set up.
Alarm relay
Input
Output
DC voltage source
Temperature sensor
Endress+Hauser
5
Identification
STIP-scan CAM74/CAS74
2
Identification
2.1
Nameplate
Compare the order code on the nameplate (transmitter and sensor) with the product structure and
your order.
Made in Germany, D-64823 Groß-Umstadt
OrderCode:
CAM74-1A3B2GB
SerialNo.:
A4000105S00
Mains:115/230VAC
50/60Hz 130VA
Channels:
Outputs:
Digital output:
Contacts:
Ambient temp.
IP65
1xSTIP-scan
4x0/4-20mA,Sensor1
ProfibusDP
7xRelais,Sensor1
-10...+40°C
140234-4A
STIP-scan
a0010013
Fig. 1:
Nameplate of transmitter (example)
Made in Germany, D-64823 Groß-Umstadt
STIP-scan NC+
OrderCode:
CAS74-SP1B
SerialNo.:
65ABBF01065
24V DC
Ip67
0 ... +30°C (+32 ... 86°F)
0 ... 0.3 bar (4.4 psi)
30VA
a0011530
Fig. 2:
6
Nameplate of sensor (example)
Endress+Hauser
STIP-scan CAM74/CAS74
Identification
2.1.1
Product structure for transmitter CAM74
Sensor input
1
2
1x STIP-scan
2x STIP-scan
Power supply
A
B
115/230 V; 50/60 Hz
100 to 250 V; 50 to 60 Hz
Analog output
1
2
3
4
5
Not selected
2x 0/4 to 20 mA; sensor 1 (not with CAM74-2******)
4x 0/4 to 20 mA; sensor 1 (not with CAM74-2******)
4x 0/4 to 20 mA; sensor 1 + 2 (not with CAM74-1******)
8x 0/4 to 20 mA; sensor 1 + 2 (not with CAM74-1******)
Digital output
A
B
Not selected
PROFIBUS DP (not with CAM74-2******)
Additional contacts
1
2
3
Not selected
7x relay; sensor 1 (not with CAM74-2******)
14x relay; sensor 1 + 2 (not with CAM74-1******)
Display
G
H
Stainless steel version, Graphical, touch screen PC
Aluminum version, Graphical, touch screen PC
Accessories
A
B
C
D
E
F
G
H
CAM74-
Not selected
1 x sensor wall mounting + transmitter wall mounting
2 x sensor wall mounting + transmitter wall mounting
1 x sensor stand + transmitter wall mounting
2 x sensor stand + transmitter wall mounting
1 x stand for sensor and transmitter
1 x stand for sensor + 1 x stand for sensor and transmitter
Weather protection cover + 1 x stand for sensor and transmitter (not with CAM74-*A*****)
complete order code
2.1.2
Product structure for sensor CAS74
Measuring parameter
NI
NS
SA
SP
Nitrate
Nitrate + SAC + COD/BOD/TOC equivalent
SAC + COD/TOC equivalent
Spectrum + Nitrate + SAC + COD/BOD/TOC equivalent
Sensor holder
0
1
2
3
4
5
6
Not selected (replacement sensor)
0.5 m + 5 m (1.64 ft + 16.4 ft) cable
0.5 m + 10 m (1.64 ft + 32.8 ft) cable
0.5 m + 20 m (1.64 ft + 65.6 ft) cable
1.5 m + 5 m (4.92 ft + 16.4 ft) cable
1.5 m + 10 m (4.92 ft + 32.8 ft) cable
1.5 m + 20 m (4.92 ft + 65.6 ft) cable
Assembly
A
B
CAS74-
Endress+Hauser
Not selected
Flow through chamber
complete order code
7
Identification
STIP-scan CAM74/CAS74
2.2
Scope of delivery
The following are included in the scope of supply of the measuring system:
• Sensor holder
• Sensor
• Data cable
• Transmitter
• Air pump
• Operating Instructions
If you have any questions, please contact your supplier or your local sales center.
2.3
Certificates and approvals
Declaration of conformity
The product meets the requirements of the harmonized European standards. It thus complies with
the legal requirements of the EC directives.
The manufacturer confirms successful testing of the product by affixing the 4 symbol.
8
Endress+Hauser
STIP-scan CAM74/CAS74
Installation
3
Installation
3.1
Incoming acceptance, transport, storage
• Make sure the packaging is undamaged!
Inform the supplier about any damage to the packaging.
Keep the damaged packaging until the matter has been settled.
• Make sure the contents are undamaged!
Inform the supplier about damage to the contents. Keep the damaged products until the matter
has been settled.
• Check that the order is complete and agrees with your shipping documents.
• The packaging material used to store or to transport the product must provide shock protection
and humidity protection. The original packaging offers the best protection. Also, keep to the
approved ambient conditions (see "Technical data").
• If you have any questions, please contact your supplier or your local sales center.
Endress+Hauser
9
Installation
STIP-scan CAM74/CAS74
3.2
!
Installation conditions
Note!
Install the sensor at a stand with supporting arms or at a wall bracket with supporting arms. Do not
install the sensor suspended from the connection cable!
310 /
12.2
mm / inch
max. 695/
27.4
4
1850 / 72.8
3
50 / 1.97
5
2
max. 300 / 11.8
1
6
a0007391
Fig. 3:
1
2
3
4
5
6
10
Stand with supporting arms
Base plate
Stand
Transmitter CAM74 in weather protection housing
Connection cable with supporting pipe
Supporting arms
STIP-scan sensor CAS74
Endress+Hauser
STIP-scan CAM74/CAS74
Installation
3.3
Installation instructions
3.3.1
Installation with upright post
105 / 4.13
210 / 8.27
150 / 5.91
150 / 5.91
210 / 8.27
1
max. 300 / 11.8
/
14
Ø .55
0
mm / inch
120 / 4.72
8 / 0.31
2
a0007125
Fig. 4:
1
2
!
Dimensions of baseplate
Midpoint of post
Basin rim
Note!
Use M12 chemical anchors to mount the upright post on a concrete base.
To mount the upright post, proceed as follows:
Endress+Hauser
1.
Move the upright post into position. The midpoint of the post must be at a maximum distance
of 300 mm (11.8 inch) from the basin rim.
2.
Draw the anchor holes for the baseplate onto the base.
3.
Drill the anchor holes using a 14 mm drill bit.
4.
Install the four chemical anchors.
5.
Screw the baseplate on tightly.
6.
Ground the upright post using a grounding cable (≥ 4 mm2 (≥12 AWG)).
11
Installation
STIP-scan CAM74/CAS74
2
1
5
5
3
4
a0009952
Fig. 5:
1
2
3
4
5
Upright post with accessories
Upright post
Supporting arm
Back support
Sensor holder
Cable conduit
To install the accessories, proceed as follows:
1.
Attach both cable conduits (pos. 5) to the upright post.
2.
Screw two supporting arms (pos. 2 into the upright post.
When doing so, please note the following installation conditions:
– Distance between the supporting arms when a short support tube is used: 25 to 35 cm (0.82
to 1.15 ft.)
– Distance between the supporting arms when a long support tube is used: 40 to 100 cm (1.3
to 3.3 ft.)
– Maximum immersion depth of sensor: 50 cm (1.6 ft.)
3.
Attach a sensor holder with back support (pos. 3 + 4, Abb. 5) to each of the two supporting
arms.
If you install the stainless steel version of the transmitter follow the steps 4 an 5:
12
4.
Screw the two support tubes of the panel PC into the center of the upright post (M10x90).
5.
Screw the weatherproof housing into the external bores of the two support tubes (M10x40).
Endress+Hauser
STIP-scan CAM74/CAS74
Installation
If you install the aluminum version of the transmitter proceed as follows:
1.
Screw the mounting bracket and the weather protection roof (if applicable) with two screws
M8 to the upright post.
2.
Install the transmitter in the mounting bracket. At first use only the center screws on both
sides.
3.
Select the desired viewing angle (3 positions are selectable).
4.
Screw in the remaining screws.
3.3.2
Installing the sensor wall-bracket
2
mm (inch)
367.5 (14.5)
1
180 (7.09)
2
3
4
1
295 (11.6)
a0010463
Fig. 6:
1
2
3
4
Wall bracket
Anchor holes for wall bracket (mounting parts: 2x hex head screw 10x80, 2x wall plug 14x75 and 2x washer 10.5)
Mounting the sensor holder (mounting parts: 2x sensor holder, 2x back support and 8x Allen screw M6x40)
Sensor holder
Back support
Mount the wall bracket as follows:
Endress+Hauser
1.
Screw the wall bracket into the desired position.
2.
Attach the sensor holder to the two support tubes.
3.
Close the pipe openings using caps.
13
Installation
STIP-scan CAM74/CAS74
3.3.3
!
Installing the sensor
Note!
Note the maximum immersion depth of 550 mm (21.7")
To install the sensor, proceed as follows:
1. Put the sensor on the sensor stand on an even surface .
2. Open the connecting clip and remove the cap.
1
3. Place the seal in the direction of the arrow on the sealing
surface of the carrier tube.
4. Pull the connector slightly out of the carrier tube, and attach
the connector to the connecting socket of the support tube.
5. Secure the connection using the screw sleeve.
2
6. Put the support tube onto the carrier tube, and secure the
flange connection using the tension clip.
3
7. Use the clamping rings to adjust the measuring position
(immersion depth) and the service position.
4
8. Attach the support tube of the sensor to the supporting arms
of the upright post.
5
a0010006
6
Fig. 8:
Seal
a0010004
Fig. 7:
1
2
3
4
5
6
14
Sensor on connecting
line
Control line
Clamping ring
Support tube
Connecting clip
Carrier tube
Sensor housing
Endress+Hauser
STIP-scan CAM74/CAS74
Installation
3.3.4
Connecting the flow assembly
The flow assembly is already fitted to the sensor.
Install the flow assembly connections in such a way that air bubbles cannot form in the area of the
flow assembly.
1
2
3
4
5
a0010450
Fig. 9:
1
2
3
4
5
3.4
Connecting the flow assembly
Inlet ventilation
Sensor
Outlet ventilation
Outlet
Flow assembly
Post-installation check
• After mounting, check all the connections to ensure they are secure and leak-tight.
• Check all cables and hoses for damage.
• Check whether the cables are routed such that they are free from electromagnetic interference
influences.
Endress+Hauser
15
Wiring
STIP-scan CAM74/CAS74
#
4
Wiring
4.1
Electrical connection
Warning!
• The electrical connection must only be carried out by a certified electrician.
• Technical personnel must have read and understood the instructions in this manual and must
adhere to them.
• Ensure that there is no voltage at the power cable before beginning the connection work.
4.2
Connecting sensor and transmitter (stainless steel
version)
Connect the sensor and the transmitter as follows:
1.
Ground the upright post or the wall bracket.
2.
Install the connecting cable between the sensor and transmitter.
3.
If the transmitter is equipped with signal outputs, connect the analog modules.
4.
If the transmitter is equipped with relay outputs, connect the relay modules.
5.
If the transmitter is equipped with a PROFIBUS card, connect the fieldbus cable to the panel
PC.
6.
Connect the power cable to your power supply.
3
4
2
5
1
a0010229
Fig. 10:
1
2
3
4
5
16
Cable connections at the weatherproof housing (underside)
Cable bushing, large (Pg threaded connection)
Cable bushing, small (Pg threaded connection)
Connection, sensor 1
Connection, sensor 2
Power cable
Endress+Hauser
STIP-scan CAM74/CAS74
Wiring
4.2.1
Electrical connection of transmitter
RD
+
BU
-
24 V
R1 no
R1 com
R2 no
R2 com
A
GN
data +
YE
data -
R3 no
R3 com
R4 no
R4 com
RD
+
BU
-
24 V
G
R5 no
R5 com
R6 no
R6 com
B
GN
data +
YE
data -
R7 no
R7 com
R1 no
R1 com
+ Iout 0
- Iout 0
C
E
R4 no
R4 com
mA
+ Iout 1
- Iout 1
+ Iout 0
- Iout 0
R2 no
R2 com
R3 no
R3 com
+ Iout 1
- Iout 1
+ Iout 0
- Iout 0
D
mA
H
R5 no
R5 com
R6 no
R6 com
R7 no
R7 com
mA
+ Iout 1
- Iout 1
L
+ Iout 0
- Iout 0
F
mA
N
I
+ Iout 1
- Iout 1
a0007181
Fig. 11:
A
B
C
D
E
Endress+Hauser
Electrical connection of the transmitter (stainless steel and aluminum version)
Signal output 2 sensor 2
F
Sensor 1
Relays 1 to 7 sensor 1
G
Sensor 2
Relays 1 to 7 sensor 2
H
Signal output 1 sensor 1
Power supply
I
Signal output 2 sensor 1
Signal output 1 sensor 2
17
Wiring
STIP-scan CAM74/CAS74
1
4
3
2
5
a0010460
Fig. 12:
1
2
3
4
5
Interior view
TA7520 converter RS-232 / RS-485
TA 7022 analog outputs
TA7067 relay outputs
Power unit for sensor
Power unit for panel PC and modules
4.2.2
Connecting the signal outputs
11 +IOUT0
-IOUT0
-IOUT1
+IOUT1
The transmitter is equipped with a maximum of two analog modules per sensor (optional). Each
module has two signal outputs.
a0010225
Fig. 13:
Connecting the signal outputs
Connect the analog modules as follows:
1.
Guide the signal cable through the cable bushing.
2.
Guide the signal cable through the cable conduits to the appropriate analog modules.
3.
Connect the signal cables to the analog modules (s. Abb. 13).
! Note!
The terminal block can be removed to facilitate access to the locking screws.
18
Endress+Hauser
STIP-scan CAM74/CAS74
Wiring
4.2.3
Connecting the relay outputs
NO
Rl1
Rl7 COM
NO
10
Rl1 COM
NO
Rl2
Rl2 COM
NO
Rl3
NO
Rl4
Rl3 COM
Rl4 COM
Rl7
NO
Rl5
Rl6 COM
Rl6
NO 1
Rl5 COM
The transmitter can be optionally fitted with one relay module per sensor. Each module has seven
relay outputs. The relay outputs function as NC contacts.
a0010226
Fig. 14:
Connecting the relay module
Connect the relay module as follows:
1.
Guide the connecting cable through the cable bushing.
2.
Guide the connecting cable through the cable conduits to the appropriate relay module.
3.
Attach the connecting cable to the relay module (s. Abb. 14).
! Note!
The terminal block can be removed to facilitate access to the locking screws.
Endress+Hauser
19
Wiring
STIP-scan CAM74/CAS74
4.2.4
Connecting the panel PC
1
2
3
5
4
6
7
8
9
10
11 12
a0009452
Fig. 15:
1
2
3
4
5
6
Panel PC connections
PROFIBUS connection
PS/2 connection
LAN connection
COM 2
COM 1
USB
7
8
9
10
11
12
USB
USB
Fuse
Ground connection
Supply voltage
On/Off switch
The panel PC is fully wired, with the exception of the optional PROFIBUS connection.
Connect the optional PROFIBUS connection as follows:
1.
Guide the PROFIBUS cable through the Pg threaded connection.
2.
Solder the D-sub connector onto the PROFIBUS cable.
3.
Plug the D-sub connector into the socket of the PROFIBUS card (pos. 1).
5
GND
9
4
8
3
DP A
DP B
7
2
6
+5V
1
a0010448
Fig. 16:
4.3
Bus cable connection
Connecting sensor and transmitter (aluminum version)
Connect the sensor and the transmitter as follows:
20
1.
Ground the upright post or the wall bracket.
2.
Install the connecting cable between the sensor and transmitter.
3.
If the transmitter is equipped with signal outputs, connect the signal outputs.
4.
If the transmitter is equipped with relay outputs, connect the relay outputs.
5.
If the transmitter is equipped with a PROFIBUS card, connect the fieldbus cable.
6.
Connect the power cable.
Endress+Hauser
STIP-scan CAM74/CAS74
Wiring
4.3.1
Electrical connection of transmitter
RD
+
BU
-
24 V
R1 no
R1 com
R2 no
R2 com
A
GN
data +
YE
data -
R3 no
R3 com
R4 no
R4 com
RD
+
BU
-
24 V
G
R5 no
R5 com
R6 no
R6 com
B
GN
data +
YE
data -
R7 no
R7 com
R1 no
R1 com
+ Iout 0
- Iout 0
C
E
R4 no
R4 com
mA
+ Iout 1
- Iout 1
+ Iout 0
- Iout 0
R2 no
R2 com
R3 no
R3 com
+ Iout 1
- Iout 1
+ Iout 0
- Iout 0
D
mA
H
R5 no
R5 com
R6 no
R6 com
R7 no
R7 com
mA
+ Iout 1
- Iout 1
L
+ Iout 0
- Iout 0
F
mA
N
I
+ Iout 1
- Iout 1
a0007181
Fig. 17:
A
B
C
D
E
Endress+Hauser
Electrical connection of the transmitter (stainless steel and aluminum version)
Signal output 2 sensor 2
F
Sensor 1
Relays 1 to 7 sensor 1
G
Sensor 2
Relays 1 to 7 sensor 2
H
Signal output 1 sensor 1
Power supply
I
Signal output 2 sensor 1
Signal output 1 sensor 2
21
Wiring
STIP-scan CAM74/CAS74
2
1
3
6
4
7
8
9
5
10
a0011505
Fig. 18:
1
2
3
4
5
Connections of the transmitter (aluminum version))
Sensor 1
Sensor 2
Power switch
Power cable (Pg threaded connection)
PROFIBUS cable (Pg threaded connection)
1
6
7
8
9
10
3
2
Relay outputs sensor 1 (Pg threaded connection)
Signal outputs sensor 1 (Pg threaded connection)
Relay outputs sensor 2 (Pg threaded connection)
PROFIBUS cable (Pg threaded connection)
Signal outputs sensor 2 (Pg threaded connection)
4
5
Sensor 1
Sensor 2
+1
+ 4
1
7
7
1
4 ... 20 mA
4 ... 20 mA
6
7
4
1
9
8
10
11
a0011442
Fig. 19:
1
2
3
4
5
6
22
Connections inside (aluminum version)
Sensor 1
Sensor 2
Power switch
Power connection
PROFIBUS connection (optional)
without function
7
8
9
10
11
Fuse
3 x USB
2 x COM RS232
LAN
PS/2
Endress+Hauser
STIP-scan CAM74/CAS74
Wiring
4.3.2
Connecting the relay and signal outputs
The transmitter is optionally equipped with:
• Seven relay outputs per sensor (the outputs function as NC contacts).
• Maximum of four signal outputs per sensor.
Connect the outputs as follows:
1.
Guide the cables through the corresponding Pg threaded connections.
2.
Connect the cables.
4.3.3
Connecting the PROFIBUS
Connect the optional PROFIBUS connection as follows:
1.
Guide the PROFIBUS cable through the Pg threaded connection.
2.
Solder the D-sub connector onto the PROFIBUS cable.
3.
Plug the D-sub connector into the socket of the PROFIBUS card.
5
GND
9
4
8
3
DP A
DP B
7
2
6
+5V
1
a0010448
Fig. 20:
4.3.4
Bus cable connection
Connecting of the power
Connect the power cable as follows:
Endress+Hauser
1.
Guide the power cable through the corresponding Pg threaded connection.
2.
Connect the power cable to the delivered female plug.
3.
Put the female plug into the socket.
23
Wiring
STIP-scan CAM74/CAS74
4.4
Post-connection check
When electrical connection is complete, perform the following checks:
Device status and specifications
Notes
Is there any damage to the exterior of the transmitters and cables?
Visual inspection
Does the mains voltage correspond to the nameplate data?
Electrical connection
Notes
Are the assembled cables strain-relieved?
Is the cable conduit free from loops and crossovers?
Visual inspection
Are all lines connected correctly in accordance with the wiring diagram?
Are all screw terminals securely tightened?
Are all cable entries assembled, secured tightly and sealed?
Is the upright post grounded?
24
Grounding is performed onsite.
Endress+Hauser
STIP-scan CAM74/CAS74
Operation
5
Operation
5.1
Human interface
1
2
3
4
5
6
a0010441
Fig. 21:
1
2
3
4
5
6
Operating elements of the panel PC
On/Off switch
Power and error LEDs
Brightness plus
Brightness minus
Right mouse button
Display and hide keyboard
The measuring system is operated using the PC touchscreen.
a0010002
Fig. 22:
Touchscreen with keyboard
You can display the keyboard using the keyboard button (to the right of the touchscreen).
Endress+Hauser
25
Operation
STIP-scan CAM74/CAS74
5.2
Onsite operation
5.2.1
MEASUREMENT menu
a0010452-en
Fig. 23:
MEASUREMENT menu
START MEASUREMENT
In the MEASUREMENT menu, select the START MEASUREMENT menu option.
The measuring system starts measurement, opens two additional windows (TRACE and
MEASURED VALUES) and generates the following message: SENSOR IN NORMAL
OPERATION
TRACE
a0009454-en
Fig. 24:
TRACE window
This windows shows the traces for the selected parameters. Details of the traces are described in
the "Traces" section.
26
Endress+Hauser
STIP-scan CAM74/CAS74
Operation
a0009455-en
Fig. 25:
MEASURED VALUES window
Stop measurement in the MEASUREMENT menu by selecting the STOP MEASUREMENT
menu option.
!
Note!
During measurement, the TEST menu and the menu options KHP VALIDATION, NOx
VALIDATION and TAKE SAMPLE SPECTRA are deactivated.
CURRENT PARAMETERS
In the MEASUREMENT menu, select the PARAMETER menu option.
a0009488-en
Fig. 26:
PARAMETERS window
Please note that:
• You can select only one parameter at a time for the parameters TOC, COD and BOD.
• You can activate SPECIAL PARAMETER only for sensor version SP (can be specified on
customer demand).
• CR stands for contamination rate.
Select the relevant parameters, and record your selection using SAVE.
Using the REJECT button, you can delete your settings and restore the previous settings.
Endress+Hauser
27
Operation
STIP-scan CAM74/CAS74
MONITORING SITE
In the MEASUREMENT menu, select the MONITORING SITE menu option.
a0009486-en
Fig. 27:
!
MONITORING SITE window
Note!
The measuring point must be specified for the purpose of plausibility testing.
Select the relevant measuring point and record your selection using SAVE.
Using the REJECT button, you can delete your settings and restore the previous settings.
28
Endress+Hauser
STIP-scan CAM74/CAS74
Operation
5.2.2
TRACES menu
a0010453-en
Fig. 28:
TRACES menu
CURRENT TRACE
This menu option displays the current trace. The trace of the scaled parameter is displayed in bold.
In the TRACES menu, select the CURRENT TRACE menu option.
2
3
1
4
TIME
= 11:20
NO3-N = 14.15 mg/l
CSB
= 7.93 mg/l
SAK
= 36.51 1/m
5
6
a0009534-en
Fig. 29:
1
2
3
4
5
6
Endress+Hauser
Current trace
The Y axis is scaled for this parameter.
You can access the daily log here.
Date and time display
Selected parameters
Trace
Window with measured values
29
Operation
STIP-scan CAM74/CAS74
Scaling
Enter the scaling as follows:
1.
Click on the parameter (pos. 1, Fig. 29).
The SCALE window opens.
1
2
a0009498-en
Fig. 30:
1
2
30
SCALE window
Parameter
Maximum value of Y axis
2.
Select the desired parameter.
3.
Enter the maximum value of the Y axis.
4.
Confirm your entries using SAVE.
Endress+Hauser
STIP-scan CAM74/CAS74
Operation
Protocol
Click on PROTOCOL (pos. 2, Fig. 29), to access the daily log.
1
2
3
a0009535-en
Fig. 31:
1
2
3
Daily log
Time
Message
Activation
The daily log lists the status messages for the sensor and the warnings and error messages.
Date and time
Date and time are displayed in the menu bar (pos. 3, Fig. 29).
Parameters and trace
The selected measurement parameters are displayed in pos. 4, Fig. 29. The color is used to identify
the associated trace (pos. 5, Fig. 29).
Measured value window
Click within the diagram along the X axis. The measured values are then displayed in a window
(pos. 6, Fig. 29) according to time.
Endress+Hauser
31
Operation
STIP-scan CAM74/CAS74
SINGLE MEASUREMENT
This menu option shows you different spectra as a function of wavelength. This menu option is
available only in the CAS74-SPxx version.
In the TRACES menu, select the SINGLE MEASUREMENT (SPECTRUM) menu option.
a0009533-en
Fig. 32:
Single measurement (spectrum)
Identifying the spectra:
Red line
Blue line
Green line
Reference intensity with distilled water
Intensity of current measurement
Measured absorption spectrum
The following settings are possible:
• Display several measurements in one diagram.
Activate OVERLAP CURVES (lower left-hand corner).
• Adjust wavelength range.
Select the desired maximum wavelength from the picklist (maximum value of X axis).
32
Endress+Hauser
STIP-scan CAM74/CAS74
Operation
SHOW HISTORICAL TRACES
You can use this menu option to page through the traces.
In the TRACES menu, select the SHOW HISTORICAL TRACES menu option.
2
1
3
4
7
6
5
a0010042-en
Fig. 33:
1
2
3
4
5
6
7
Show historical traces
Parameter whose Y axis scale is active
Time window
Protocol displays the daily log
Selected parameters
Saves the measured values in a csv file
Date display
Zoom function (scaling of the time axis)
Time window
Move the time window (pos. 2, Fig. 33) to where you want to look at the trace in greater detail.
Then click on the 3 HOURS button. The trace is displayed for the selected time period. One click
on the 24 HOURS button brings you back.
Endress+Hauser
33
Operation
STIP-scan CAM74/CAS74
Daily log
Click on PROTOCOL (pos. 3, Fig. 33) to view the daily log.
1
3
2
a0009535-en
Fig. 34:
1
2
3
Daily log
Time
Message
Activation
Scaling of Y axis
Scaling of the Y axis is determined by a parameter (pos. 1, Fig. 33). In this screenshot it is NOx-N.
To change the scaling, click on the parameter displayed. The following window appears:
1
2
a0009498-en
Fig. 35:
1
2
34
Scaling of Y axis
Parameter
Maximum value of Y axis
Endress+Hauser
STIP-scan CAM74/CAS74
Operation
Selected parameters
The parameters selected for measurement are displayed here (pos. 4, Fig. 33). Each parameter is
assigned its own color for trace identification.
Save
Click on the SAVE button (pos. 5, Fig. 33) to save the measured values for a selected day in a csv file.
Date display
The date display (pos. 6, Fig. 33) shows you when the trace was created. You can use the relevant
arrow keys to page back and forth.
Zoom function
You can use this button (pos. 7, Fig. 33) to choose between the 24-hour display and the 3-hour
display.
Endress+Hauser
35
Operation
STIP-scan CAM74/CAS74
SLUDGE SETTLING CURVE
This menu option shows you the sludge settling curve.
In the TRACES menu, select the SLUDGE SETTLING CURVE menu option.
a0010041-en
Fig. 36:
Sludge settling curve
The following settings are possible:
• Display several measurements in one diagram.
Activate OVERLAP CURVES (lower left-hand corner).
• Display slope of settling curve.
Activate DISPLAY SLOPE.
• Display absorption spectrum.
Activate DISPLAY SPECTRUM.
!
Note!
The DISPLAY SLOPE and DISPLAY SPECTRUM functions are available only in the CAS74-SPxx
version.
MEASURED VALUES
This menu option shows you the measured values in tabular form.
In the TRACES menu, select the MEASURED VALUES menu option.
a0009455-en
Fig. 37:
36
Measured values in tabular form
Endress+Hauser
STIP-scan CAM74/CAS74
Operation
TRACE POLLUTION
This menu option shows you the pollution level of the cell on air.
In the TRACES menu, select the TRACE POLLUTION menu option.
a0010158-en
Fig. 38:
Trace pollution
If the CR parameter is activated, the contamination rate is measured automatically every 6 hours:
CR 0
CR 1
CR 2
CR 3
Endress+Hauser
00:00 hours
06:00 hours
12:00 hours
18:00 hours
37
Operation
STIP-scan CAM74/CAS74
5.2.3
CALIBRATION menu
a0010454-en
Fig. 39:
CALIBRATION menu
Calibration is used to adapt the measuring system to the conditions of the measuring point.
• Do a separate calibration for each of the following parameters: NOx-N, CODeq or TOCeq or
BODeq, SV and TS.
• The parameters SAC, SI and ATU do not need to be calibrated.
Preparing for calibration
For calibration you need laboratory values from samples and the corresponding measured values
from the measuring system.
These values are obtained as follows:
1.
Start measurement and allow the measuring system to measure for an entire day.
2.
After one day, take a look at the traces for the day, and determine for each parameter the time
at which the maximum and minimum measured values occur. The difference between
maximum and minimum should be at least 30% of the maximum measured value (not for the
parameters SV and TS).
3.
On the following day, take three samples for each parameter at the times calculated (minimum
and maximum). Note the associated measured values.
4.
Before analyzing the samples in the laboratory, the samples must be pretreated as follows:
Measuring point
Pretreatment
Inlet
Settling of suspended matter
Sludge activation
Filtering out of suspended matter
Outlet
No pretreatment required
5.
Perform the reference measurements for the samples in the laboratory.
Using the measured values from the measuring system and the measured values from the laboratory,
you can now perform the TARGET/ACTUAL ADJUSTMENT.
38
Endress+Hauser
STIP-scan CAM74/CAS74
Operation
TARGET/ACTUAL ADJUSTMENT
This menu option is used to calibrate the measuring system.
In the CALIBRATION menu, select the TARGET/ACTUAL ADJUSTMENT menu option.
a0009499-en
Fig. 40:
Calibration menu
Carry out the calibration as follows:
1.
Select the calibration parameter (e.g. NOx-N).
2.
In the "STIP-scan" column, enter the highest and the lowest value.
3.
In each of the columns "Lab. 1 - 3", enter the value calculated in the laboratory.
4.
Activate entries using the SAVE button.
The system uses the values entered to calculate the calibration constants (slope and zero point shift
of the calibration lines).
The measuring system is thus calibrated for the selected parameter.
Using the RESET CALIBRATION button, you can reset the calibration constants to the factory
setting.
ENTER MEASURING FUNCTION
With this menu option you enter a quadratic measuring function. This quadratic measuring function
overwrites the TARGET/ACTUAL ADJUSTMENT. The RESET FUNCTION button is used to
reactivate the TARGET/ACTUAL ADJUSTMENT.
If you require the quadratic measuring function, please contact E+H Service. E+H Service calculates
the data for your measuring function.
In the CALIBRATION menu, select the ENTER MEASURING FUNCTION menu option.
a0009495-en
Fig. 41:
Enter measuring function
Enter the measuring function as follows:
Endress+Hauser
1.
Select the parameter (e.g. NOx-N).
2.
Enter the three values for the measuring function.
3.
Activate entries using the SAVE button.
39
Operation
STIP-scan CAM74/CAS74
KHP VALIDATION
This menu option is used to check the measuring system for the carbon parameters using a KHP
standard solution. To do this, you need a KHP solution with a concentration of 50 mg/l KHP.
For KHP validation, proceed as follows:
1.
In the CALIBRATION menu, select the KHP VALIDATION menu option.
The measuring system empties the measuring cell. After the measuring cell has been emptied,
you are requested to immerse the sensor in the KHP solution.
2.
Remove the sensor from the wastewater.
3.
Thoroughly clean the exterior of the sensor.
4.
Place the sensor in the KHP solution.
5.
Use the OK button to confirm that the sensor is in the KHP solution.
The quartz cell is rinsed times in KHP solution before starting measurement. Once measurement
has been completed, the value calculated by the measuring system is displayed.
!
Note!
If the value calculated deviates from the set point of the KHP solution by more than 10%, please
contact Endress+Hauser Service.
NOx VALIDATION
This menu option is used to check the measuring system for the nitrogen parameters using an NOx
standard solution. To do this, you need an NOx solution with a concentration of 10 mg/l NO3-N.
For NOx validation, proceed as follows:
1.
In the CALIBRATION menu, select the NOx VALIDATION menu option.
The measuring system empties the measuring cell. After the measuring cell has been emptied,
you are requested to immerse the sensor in the NOx solution.
2.
Remove the sensor from the wastewater.
3.
Thoroughly clean the exterior of the sensor.
4.
Place the sensor in the NOx solution.
5.
Use the OK button to confirm that the sensor is in the NOx solution.
The quartz cell is rinsed times in NOx solution before starting measurement. Once measurement
has been completed, the value calculated by the measuring system is displayed.
!
40
Note!
If the value calculated deviates from the set point of the NOx solution by more than 10%, please
contact Endress+Hauser Service.
Endress+Hauser
STIP-scan CAM74/CAS74
Operation
TAKE SAMPLE SPECTRA
This menu option is used to record liquids in the range of any wavelengths. This menu option is
available only in the CAS74-SPxx version.
In the CALIBRATION menu, select the TAKE SAMPLE SPECTRA menu option.
a0009496-en
Fig. 42:
TAKE SAMPLE SPECTRA window
Enter the following data:
Entry field
Description
SAMPLE NO.
Enter the number of your sample.
CONCENTRATION (mg/l)
Enter the concentration in mg/l.
NUMBER OF FLUSHES
Enter how many flushes are necessary prior to measurement.
SPECTRUM BETWEEN (nm)
Enter the wavelength at which absorption measurement is to commence.
AND (nm)
Enter the wavelength up to which absorption measurement is to take place.
FILE NAME
Enter the file name under which the measured values are to be saved.
Start measurement using the START MEASUREMENT button.
The measurement results are displayed in the right-hand window and in the trace window which
opens.
The measured values for absorption and intensity are saved to the following files:
• Measured values for absorption: FILENAME.p_a
• Measured values for intensity: FILENAME.p_i
The files can be viewed using Microsoft Excel.
Endress+Hauser
41
Operation
STIP-scan CAM74/CAS74
CAL.CONSTANTS
This menu option displays the calibration constants.
In the CALIBRATION menu, select the CAL.CONSTANTS menu option.
1
2
a0009492-en
Fig. 43:
1
2
42
CAL.CONSTANTS window (before calibration was performed)
Calibration constants for TARGET/ACTUAL ADJUSTMENT (linear)
Calibration constants for ENTER MEASURING FUNCTION (quadratic)
Endress+Hauser
STIP-scan CAM74/CAS74
Operation
5.2.4
SETTINGS menu
a0010455-en
Fig. 44:
SETTINGS menu
NAMING THE MEASURING POINT
In this window you can enter a name for the measuring point. This name appears in the title bar of
all windows.
In the SETTINGS menu, select the MEASURING POINT menu option
a0009481-en
Fig. 45:
MEASURING POINT window
Record the name using SAVE.
Endress+Hauser
43
Operation
STIP-scan CAM74/CAS74
CONFIGURATION OF THE MEASURING POINT
In this window you configure the measuring point.
In the SETTINGS menu, select the CONFIGURATION OF THE MEASURING POINT. menu
option
a0009485-en
Fig. 46:
44
CONFIGURATION OF THE MEASURING POINT window
Parameter
Range of
adjustment
Description
Max. settling time
(seconds)
50 to 3000
Enter the maximum time for the sludge to settle. If this time is
exceeded, the warning "No sludge settling" is activated.
(Can be configured at measuring point: activated sludge basin)
Abort from (% first
value)
10 to 70
Here you enter when the observation of the sludge settling is to be
suspended and measurement of the slope threshold is to commence
(enter as a percentage of the first value).
(Can be configured at measuring point: activated sludge basin)
Slope threshold
(ppm)
1 to 5
Enter the limit value of the slope for the sludge settling curve. Once the
limit value has been reached, analysis of the sludge settling curve is
stopped. Once the measurement delay has elapsed, parameter
measurement starts.
(Can be configured at measuring point: activated sludge basin)
Delay measurement
(seconds)
5 to 3000
Enter the delay between sludge settling and the start of measurement.
(Configurable at the measuring point: activated sludge basin and inlet)
Absorption limit
15 to 150
Limit value setting for sludge settling; a higher value indicates that
settling is quick.
(Can be configured at measuring point: sludge activation basin)
Measurements per
hour
1, 2, 3, 4, 6, 12,
30
Enter the number of measuring cycles per hour.
(Configurable at measuring point: activated sludge basin, inlet and
outlet)
Endress+Hauser
STIP-scan CAM74/CAS74
Operation
SMOOTHING
In this window you can configure the smoothing of the traces.
In the SETTINGS menu, select the SMOOTHING menu option.
a0009483-en
Fig. 47:
SMOOTHING window
You can make the following settings:
• In the THRESHOLD SMOOTHING(%) column, enter the percentage change in measured value
from which the traces are to be smoothed.
• In the THRESHOLD STEP(%) column, enter the percentage change in the case of three
consecutive measured values, from which a jump in the trace is to be displayed.
• In the third column, activate the desired parameters.
Example:
• In the case of the NOx-N parameter (s. Abb. 47), the trace is smoothed as soon as a change of ≥4
% occurs between two measured values, and
• a jump appears in the trace as soon as a change of ≥10 % occurs in three consecutive measured
values.
Activate entries using SAVE.
Endress+Hauser
45
Operation
STIP-scan CAM74/CAS74
ANALOG OUTPUTS
In this window you can configure the analog outputs.
In the SETTINGS menu, select the ANALOG OUTPUTS. menu option.
The measuring system is equipped with a maximum of two analog modules per sensor. Each
module has two analog outputs. These two modules are not interchangeable (bus address).
If no analog module is connected, this menu option cannot be accessed.
1
3
2
4
a0009490-en
Fig. 48:
1
2
3
4
ANALOG OUTPUTS window
Selection field 0 - 20 mA or 4 - 20 mA
Activation field for scaling
Entry field for scaling
Entry field for current strength (for test purposes only)
To configure the analog outputs, proceed as follows:
1.
In the OP.MODE selection field (pos. 1, Fig. 48), select the desired current range.
2.
In the selection fields SIGNAL 1-4, select the desired parameters.
3.
If you wish to scale an analog output, activate the corresponding output (pos. 2, Fig. 48).
4.
In the entry field (pos. 3, Fig. 48), enter the desired maximum value for the corresponding
output.
5.
Activate entries using the SAVE button.
If scaling is not activated, the factory setting will be used for the maximum value. The REJECT
button is used to activate the previous settings.
Functional test for analog outputs
To carry out a functional test on the analog outputs, proceed as follows:
46
1.
Suspend operation.
2.
Connect an amperemeter to the relevant analog output.
3.
In the relevant SIGNAL 1-4 selection field, select the TEST parameter.
4.
In the TEST entry field (pos. 4, Fig. 48), enter a current value (max. 20 mA).
5.
Activate the test using the OUTPUT button.
6.
Compare the measurement result with the current value entered.
Endress+Hauser
STIP-scan CAM74/CAS74
Operation
RELAY CONTACTS
In this window you can configure the relay contacts.
If no digital module is connected, this menu option cannot be accessed.
In the SETTINGS menu, select the RELAY CONTACTS menu option.
• There are seven relay contacts available.
• The relay contacts function as NC contacts during operation.
• Relay 1 always transmits the measurement status.
• Relay 2 always transmits the leakage signal.
• Relay 3 always transmits the alarm signal from the lamp/spectrometer.
• Relays 4 - 7 can be allocated individually.
For relays 4-7, the following selections are possible:
Signal
GENERIC ALARM 1
Description
The relay opens if one of the following errors occur:
•
•
•
•
•
•
GENERIC ALARM 2
The relay opens if one of the following warnings occurs:
•
•
•
•
Endress+Hauser
LEAKAGE
NO LIGHT SIGNAL
NETWORK FAILURE
OUTPUT FAILURE (error & warning)
STEPPER FAILURE
AIR IN MEASURING CELL (error only)
NO SLUDGE SETTLING
ABSORPTION TOO HIGH
NITRATE VALUE TOO HIGH
AIR IN MEASURING CELL (warning only)
NETWORK FAILURE
The relay opens if communication between controller and sensor is suspended.
OUTPUT FAILURE
The relay opens if there is a failure in the data output via the relay contacts or
the analog outputs.
STEPPER FAILURE
The relay opens if there is a failure in the stepper.
AIR IN MEASURING CELL
The relay opens if air is detected in the measuring cell.
NITRATE VALUE TOO HIGH
The relay opens if the measuring range is exceeded when measuring nitrate.
COD/TOC/SAC VALUE TOO
HIGH
The relay opens if the measuring range is exceeded when measuring COD,
BOD, TOC or SAC.
REFERENCE MEASUREMENT
The relay opens when this operating status is reached.
NOx VALIDATION
The relay opens when this operating status is reached.
KHP VALIDATION
The relay opens when this operating status is reached.
MEASURING OF
CONTAMINATION
The relay opens when this operating status is reached.
MEAS. CELL CONTAMINATED
The relay opens when the contamination measurement exceeds this threshold.
MEAS. CELL HEAVILY
CONTAMINATED
The relay opens when the contamination measurement exceeds this threshold.
47
Operation
STIP-scan CAM74/CAS74
a0009497-en
Fig. 49:
RELAY CONTACTS window
To configure the relay contacts 4-7, proceed as follows:
1.
In the RELAY 4 selection field, select the data to be transmitted via this relay.
2.
Repeat this for RELAY 5 - 7.
3.
Activate entries using the SAVE button.
Functional test for relays
To carry out a functional test on the relays, proceed as follows:
48
1.
In the TEST selection field, select the relevant relay.
2.
Confirm the relay using the CLOSE and OPEN buttons.
3.
Test the relay function using a continuity tester or ohmmeter.
Endress+Hauser
STIP-scan CAM74/CAS74
Operation
PASSWORD
In this window you can activate the password function.
In the SETTINGS menu, select the PASSWORD menu option.
a0010003-en
Fig. 50:
PASSWORD window
Using this password function you can protect the
• CALIBRATION
• SETTINGS
menus from unauthorized access.
!
Note!
Once activated, it is no longer possible to deactivate the password function! Deactivation of the
password function can be performed only by Endress+Hauser Service.
Activate the password function as follows:
1.
For initial activation, enter the manufacturer password. The password can be found on the
leaflet enclosed with your device.
2.
Confirm the password using the ENTER button.
3.
In the next window enter the new password.
4.
Confirm the new password using the ENTER button.
5.
In the next window enter the new password again.
6.
Confirm again using the ENTER button.
The password function is now activated. Activation is indicated by a red dot in the upper left-hand
corner of the main window.
Endress+Hauser
49
Operation
STIP-scan CAM74/CAS74
INTERFACE
In the SETTINGS menu, select the INTERFACE menu option.
In this window you can select the interface:
a0009484-en
Fig. 51:
Selecting the interface
• Check which interface the sensor is connected to.
• For the first sensor select the COM 1 interface and confirm with OK.
• For the second sensor select the COM 2 interface and confirm with OK.
50
Endress+Hauser
STIP-scan CAM74/CAS74
Operation
ADDITIONAL OUTPUTS
In the SETTINGS menu, select the ADDITIONAL OUTPUTS menu option.
In this window you can activate the following options:
a0009512-en
Fig. 52:
ADDITIONAL OUTPUTS window
Option
Function
FIELDBUS
Activate the fieldbus connection.
FB Adr
Enter the fieldbus address here.
SAVE SLUDGE CURVES
Activate saving of sludge data to a csv file.
SHOW PROGRESS BAR
Activate the following progress bars in the main window:
• Next measurement
• Sludge settling time
• Settling time
Endress+Hauser
SAVE CURRENT SPECTRUM
Activate saving of the spectrum to a csv file. This function requires a lot
of memory!
SAVE DAILY DATA
Activate saving the measured values for an entire day - with the
exception of the sludge data - to a csv file.
SAVE RAW DATA
Activate saving of raw data for carbon and nitrate calculation without
calibration function. These values are used to determine a measuring
function. The data are saved in csv format.
51
Operation
STIP-scan CAM74/CAS74
Save sludge curves
In the ADDITIONAL OUTPUTS window, activate the SAVE SLUDGE CURVE option.
All data for the sludge settling curve (SDA) are saved in csv files on the hard disk of the panel PC.
The following files are created:
• File containing the sludge measurement values for the time hh-mm
The SDA-hh-mm.csv files are created for each sludge measurement at the actual time in question.
They are saved to the following location:
C:\STIP-scan\SDAJJJJ\SDAJJJJ-MM\SDAJJJJ-MM-TT\SDA_hh-mm.csv.
• File containing the sludge measurement values for the day JJJJ-MM-TT
All the sludge data for one day are also saved in one single file. They are saved to the following
location: C:\STIP-scan\SDAJJJJ\SDAJJJJ-MM\SDAJJJJ-MM-TT\SDA_JJJJ-MM-TT.csv.
Explanation of file names:
SDA
JJJJ
mm
TT
hh
mm
sludge data
four-digit year
two-digit month
two-digit day
two-digit hour
two-digit minutes
Save current spectrum
In the ADDITIONAL OUTPUTS window, activate the SAVE CURRENT SPECTRUM option.
The spectrum for the current measurement and the spectra for KHP and NOx validation are saved
to a file which can be viewed using Microsoft Excel.
• The file containing the current spectrum is named as follows: AKT_SPEK_hh-mm.MES.
Location: C:\STIP-scan\AKT_SPEKJJJJ\JJJJ-MM\JJJJ-MM-TT\AKT_SPEK_hh-mm.MES
• The file containing the NOx validation spectrum is named as follows: AKT_SPEK_hh-mm.NOX.
Location: C:\STIP-scan\AKT_SPEK_hh-mm.NOX
• The file containing the KHP validation spectrum is named as follows: AKT_SPEK_hh-mm.KHP.
Location: C:\STIP-scan\AKT_SPEK_hh-mm.KHP
Processing of data using Microsoft Excel
You can convert the csv files to an Excel table and work on them there.
To do so, proceed as follows:
!
1.
Copy the csv files from the hard disk of the panel PC to a USB stick.
2.
Copy the csv files from the USB stick to your "Excel computer".
3.
Start Microsoft Excel.
4.
Open the desired csv file using the File > Open menu.
Note!
The csv files have the following factory settings:
• lists separated by a semicolon
• no 1000s separator
• comma used as decimal separator
In the SETTINGS menu, select the CSV SETTINGS menu option, if you wish to view or change
the settings
52
Endress+Hauser
STIP-scan CAM74/CAS74
Operation
Save daily data
In the ADDITIONAL OUTPUTS menu, activate the SAVE DAILY DATA option.
The measured values for all parameters are saved to the hard disk of the panel PC in a day file called
DAJJJJMMTT.csv.
The day files are stored in the relevant month folder MDAJJJJMM.
Location: C:\STIP-scan\MDAJJJJMM\DAJJJJMMTT.csv
Explanation of file names:
MDA
DA
JJJJ
MM
TT
month data
data
four-digit year
two-digit month
two-digit day
CSV SETTINGS
In the SETTINGS menu, select the CSV SETTINGS menu option.
In this window you can determine the following:
• Date format
• Column separator
• Decimal separator
a0009482-en
Fig. 53:
CSV SETTINGS window
Activate the settings using SAVE.
Endress+Hauser
53
Operation
STIP-scan CAM74/CAS74
5.2.5
TEST menu
a0010456-en
Fig. 54:
TEST menu
SPECTROMETER
This menu option is used to test the function of the spectrometer. In the TEST menu, select the
SPECTROMETER menu option.
a0009500-en
Fig. 55:
SPECTROMETER window
• Start single measurement using the SINGLE MEASUREMENT button. The SPECTRUM
window opens automatically.
• Start continuous measurement using the CONTINUOUS MEASUREMENT button. The button
now appears in red. The SPECTRUM window opens automatically. If you activate the
OVERLAP CURVES function in the SPECTRUM window, the measurement curves for each
measurement are shown in this one graph. You can stop continuous measurement by pressing the
CONTINUOUS MEASUREMENT button again.
54
Endress+Hauser
STIP-scan CAM74/CAS74
Operation
STEPPER
In the TEST menu, select the STEPPER menu option.
a0009501-en
Fig. 56:
STEPPER window
Button
Function
STEPPER UPWARDS
The piston travels all the way to the top. The sensor is filled completely with
medium.
STEPPER DOWNWARDS
The piston travels all the way to the bottom. The sensor is emptied completely.
FLUSH 5 TIMES
The piston travels up and down 5 times. Use this function to clean or flush the
sensor using distilled water or cleaning solution.
CONTINUOUS FLUSH
The piston travels up and down continuously. To stop continuous flushing, press
the CONTINUOUS FLUSH button again. While this function is active, the
button is highlighted in red.
AIR PRESSURE
In the TEST menu, select the AIR PRESSURE menu option.
1
3
2
4
a0009494-en
Fig. 57:
1
2
3
4
Endress+Hauser
AIR PRESSURE window
MEMO - displays the air pressure value saved
Button for saving the current air pressure
CURRENT - displays the current air pressure
Button to close the AIR PRESSURE window
55
Operation
STIP-scan CAM74/CAS74
MEASURE CONTAMINATION RATE
This menu option is used to measure the contamination level of the sensor.
In the TEST menu, select the MEASURE CONTAMINATION RATE menu option.
Measurement takes several minutes and cannot be interrupted.
a0010250-en
Fig. 58:
MEASURE CONTAMINATION RATE, result
If the CR parameter is activated in the MEASUREMENT > CURRENT PARAMETERS menu,
the contamination rate is measured automatically every 6 hours (0:00, 6:00, 12:00 and 18:00
hours).
The contamination rate is also indicated by a traffic light. If the first threshold value is exceeded, the
traffic light switches to yellow. If the second threshold value is exceeded, the traffic light switches
to red. Measurement continues in both cases.
You can configure one relay output for threshold value overshoot.
56
Endress+Hauser
STIP-scan CAM74/CAS74
Operation
TEST SENSOR
This menu option is used to test the functional capability of the entire system.
In the TEST menu, select the TEST SENSOR menu option.
a0009453-en
Fig. 59:
TEST SENSOR, result
The following information is listed:
• The serial number of the sensor with the associated module type
• Temperature of the sensor interior
• Air pressure when the sensor is open (no overpressure in sensor interior)
• Current air pressure in the sensor (after pressurization)
• Message from leakage sensor (here: NO LEAKAGE)
• Serial number and status of spectrometer
• Status of reference data
• Status of stepper
• Testing the analog signal outputs (display of module series)
• Testing of relay contacts (display of module series)
• End of system test
This system test is also carried out automatically at each system startup.
Endress+Hauser
57
Operation
STIP-scan CAM74/CAS74
5.2.6
LANGUAGE menu
Select the LANGUAGE menu.
a0010458-en
Fig. 60:
LANGUAGE menu
Click on the desired language.
58
Endress+Hauser
STIP-scan CAM74/CAS74
#
Commissioning
6
Commissioning
6.1
Installation and functional check
Warning!
• Check all connections for correctness.
• Make sure that the supply voltage is identical to the voltage indicated on the nameplate!
6.2
Quick setup
After power-up you must make a few settings to configure the most important transmitter functions
which are necessary to ensure correct measurement.
1.
In the LANGUAGE menu, select the desired language.
2.
In the SETTINGS menu, select the INTERFACE menu option. Select COM1 for the first
sensor and COM2 (if available) for the second sensor.
3.
Carry out a leakage test (see "Maintenance" section).
4.
In the MEASUREMENT menu, select the CURRENT PARAMETERS menu option.
Activate the parameters to be measured.
5.
In the MEASUREMENT menu, select the MONITORING SITE menu option. Enter the
location of the sensor.
6.
Configure the analog outputs (if available).
In the SETTINGS menu, select the ANALOG OUTPUTS menu option.
Select the mode of operation and the parameters.
7.
Configure the relay outputs (if available).
In the SETTINGS menu, select the RELAY OUTPUTS menu option.
Relay1 to Relay3 are allocated as follows:
– Relay1 = Operation
– Relay2 = Leakage
– Relay3 = Lamp/spectrometer failure
Relay4 to Relay7 can be individually allocated.
8.
Perform a 10-minute stepper test (sensor hanging in water):
a. In the TEST menu, select the STEPPER menu option.
b. Click on the CONTINUOUS FLUSH button.
c. Stop the test after 10 minutes by clicking again on the CONTINUOUS FLUSH button.
9.
Endress+Hauser
Start measurement:
In the MEASUREMENT menu, select the START MEASUREMENT menu option.
Allow the system to run for an entire day.
59
Commissioning
STIP-scan CAM74/CAS74
6.3
Communication
The panel PC must be fitted with a fieldbus card (CAM74-1**B***).
For fieldbus communication with the measuring system, PROFIBUS is used with the DPV1 or DPV0
protocols. The measuring system behaves like a PROFIBUS standard slave. Any device compatible
with PROFIBUS DPV1 or DPV0 can be used as a master (e.g. SIMATIC S5 or SIMATIC S7).
The PROFIBUS partner is defined by the device database (GSD file) from Hilscher. Integration into
the PROFIBUS network takes place as follows:
• in the case of the SIMATIC S5 via the "COM-PROFIBUS" program
• in the case of the SIMATIC S7 through integration into the hardware configuration of the
SIMATIC Manager.
Configure address
In the SETTINGS menu, select the ADDITIONAL OUTPUTS menu option.
In this window you can activate the PROFIBUS connection and configure the address.
!
Note!
This setting is used to initialize the PROFIBUS device when starting the program. If the address is
changed, the PROFIBUS device is initialized again.
The factory setting for the PROFIBUS address is 123. If you are already using this address, change
the address before connecting the transmitter to your network.
60
Endress+Hauser
STIP-scan CAM74/CAS74
Commissioning
Data structure
Detailed information using the example of the Siemens S7:
Parameter
Description (1st sensor)
Description (2nd sensor)
NOx
tIOSendData[0] = byte1
tIOSendData[1] = byte2
tIOSendData[2] = byte3
tIOSendData[3] = byte4
tIOSendData[36] = byte1
tIOSendData[37] = byte2
tIOSendData[38] = byte3
tIOSendData[39] = byte4
COD or TOC or BOD
tIOSendData[4] = byte1
tIOSendData[5] = byte2
tIOSendData[6] = byte3
tIOSendData[7] = byte4
tIOSendData[40] = byte1
tIOSendData[41] = byte2
tIOSendData[42] = byte3
tIOSendData[43] = byte4
Special parameter 1
tIOSendData[8] = byte1
tIOSendData[9] = byte2
tIOSendData[10] = byte3
tIOSendData[11] = byte4
tIOSendData[44] = byte1
tIOSendData[45] = byte2
tIOSendData[46] = byte3
tIOSendData[47] = byte4
SAC
tIOSendData[12] = byte1
tIOSendData[13] = byte2
tIOSendData[14] = byte3
tIOSendData[15] = byte4
tIOSendData[48] = byte1
tIOSendData[49] = byte2
tIOSendData[50] = byte3
tIOSendData[51] = byte4
Sludge volume
tIOSendData[16] = byte1
tIOSendData[17] = byte2
tIOSendData[18] = byte3
tIOSendData[19] = byte4
tIOSendData[52] = byte1
tIOSendData[53] = byte2
tIOSendData[54] = byte3
tIOSendData[55] = byte4
Total solids
tIOSendData[20] = byte1
tIOSendData[21] = byte2
tIOSendData[22] = byte3
tIOSendData[23] = byte4
tIOSendData[56] = byte1
tIOSendData[57] = byte2
tIOSendData[58] = byte3
tIOSendData[59] = byte4
Sludge index
tIOSendData[24] = byte1
tIOSendData[25] = byte2
tIOSendData[26] = byte3
tIOSendData[27] = byte4
tIOSendData[60] = byte1
tIOSendData[61] = byte2
tIOSendData[62] = byte3
tIOSendData[63] = byte4
ATU
tIOSendData[28] = byte1
tIOSendData[29] = byte2
tIOSendData[30] = byte3
tIOSendData[31] = byte4
tIOSendData[64] = byte1
tIOSendData[65] = byte2
tIOSendData[66] = byte3
tIOSendData[67] = byte4
Statuses
tIOSendData[32] = byte1
tIOSendData[33] = byte2
tIOSendData[34] = byte3
tIOSendData[35] = byte4
tIOSendData[68] = byte1
tIOSendData[69] = byte2
tIOSendData[70] = byte3
tIOSendData[71] = byte4
The sum of the bytes transmitted is 36. For versions with two sensors the sum of bytes transmitted
is 72 (also if only one sensor is connected). Ensure that the same number of bytes is retrieved in the
control unit. In the case of the Siemens S5, the sum of the bytes transmitted is 20.
Endress+Hauser
61
Commissioning
STIP-scan CAM74/CAS74
Parameters
Description
Operating statuses
0x80000000 // in operation
0x00000001 // measuring mode
0x00000002 // ref measurement
0x00000004 // NOx validation
0x00000008 // KHP validation
0x00800000 // contamination rate measurement
Error
0x00000010 // leakage
0x00000020 // light outage
0x00000040 // network failure
0x00000080 // output failure
0x00000100 // stepper failure
0x00000200 // air in sensor
0x00000800 // no valid reference
Warnings
0x00001000 // sludge not settling
0x00002000 // absorption too high
0x00004000 // nitrate value too high
0x00008000 // air in sensor
0x00010000 // CR warning, yellow traffic light
0x00020000 // CR alarm, red traffic light
The values of the Siemens S5 are in the same order. However, when transmitting the analog values,
only 2 bytes are used instead of 4. The following addresses relocate accordingly.
In the case of the S5, the data are transmitted in the form of an "Unsigned Integer" (2 bytes). In the
case of the S7, the data are transmitted as floating-point numbers in "Real Format" in accordance
with IEEE-FP-32. The factory setting is S7 which means that 4-byte values are transmitted.
!
Note!
If you want to connect the transmitter to an S5, contact E+H Service or inform your E+H sales
representative beforehand.
Integration with SIMATIC S7
Integrate the transmitter with the SIMATIC S7 as follows::
1.
Install the GSD file.
2.
Add a new slave to your hardware configuration and assign it an address in your network.
a0010291-en
Fig. 61:
62
User interface SIMATIC S7
Endress+Hauser
STIP-scan CAM74/CAS74
Commissioning
3.
Retrieve the transmitter data. 9x4 bytes are retrieved as 4 byte input con (0x93). Address
allocation starts at 0.
a0010292-en
Fig. 62:
SIMATIC S7 user interface
The data are now available for processing in the SIMATIC S7.
Endress+Hauser
63
Maintenance
STIP-scan CAM74/CAS74
7
Maintenance
Take all necessary measures in time to ensure the operational safety and reliability of the entire
measuring system.
Depending on the application, carry out maintenance work at least once a month.
Maintenance work on the device includes the following:
• Cleaning the cell
• Visual inspection of sensor housing
• Visual inspection of inlet
• Leakage test
• Review of warning and error messages
#
Warning!
There is a risk of infection when working with wastewater!
Therefore, wear protective gloves, goggles and protective clothing.
7.1
Maintenance of complete measuring point
7.1.1
Cleaning the transmitter
Clean the front of the housing with usual commercial cleaning agents.
In accordance with DIN 42 115, the front is resistant to:
• Isopropanol
• Diluted acids (max. 3%)
• Diluted alkalis (max. 5%)
• Esters
• Hydrocarbons
• Ketones
• Household cleaners
"
Caution!
For cleaning purposes, never use:
• Concentrated mineral acids or alkalis
• Benzyl alcohol
• Methylene chloride
• High-pressure steam
7.1.2
Cleaning the sensor
Clean the cell as follows:
!
64
1.
Place the sensor in a container of distilled water.
2.
In the TEST menu, select the STEPPER menu option.
3.
Click on the FLUSH 5 TIMES button.
4.
When the flushing process has been completed, click on the DOWNWARDS button.
Note!
For dirt that is hard to remove, use a cleaning solution (depending on the type of contamination
involved use 5% hydrochloric acid or 5% sodium hydroxide). Following this you must repeat the
flushing processes using distilled water.
Endress+Hauser
STIP-scan CAM74/CAS74
Maintenance
7.1.3
Leakage test
For the leakage test, an overpressure of approx. 300 mbar (approx. 4.5 psi) is generated in the
sensor. The drop in pressure is then measured. The drop in pressure within a period of 15 minutes
must not exceed 5 mbar.
For the leakage test you will need an air pump (included in the scope of supply).
1
3
2
4
a0009494-en
Fig. 63:
1
2
3
4
AIR PRESSURE window
MEMO - recorded air pressure value
Button for recording the current air pressure
ACTUAL - current air pressure
Button - to close the window
Prepare for the leakage test as follows:
1.
Ensure that the connecting cable has been installed correctly.
2.
In the TEST menu, select the AIR PRESSURE menu option.
Carry out the leakage test as follows:
1.
Pump air into the sensor using the air pump (included in the scope of supply) until the
CURRENT value is approx. 300 mbar above the MEMO value.
2.
Save the current air pressure. MEMO and CURRENT now both display the same value.
3.
After 15 minutes check the CURRENT value.
Drop in pressure < 5 mbar
Sensor is airtight.
Drop in pressure < 5 mbar
Sensor is not airtight.
If the sensor is not airtight, contact E+H Service.
Endress+Hauser
65
Accessories
STIP-scan CAM74/CAS74
8
Accessories
Supporting arms
• complete set 700 mm; supplement to 71013968; stainless steel 1.4301 (AISI 304)
• order no. 71013964
Wall mounting for transmitter and one sensor
• stainless steel 1.4301 (AISI 304)
• order no. 71013961
Mounting material
• mounting material for wall mounting of the transmitter (stainless steel version)
• order no. 71013971
Stand and mounting material for transmitter and one sensor
• stainless steel 1.4301 (AISI 304)
• order no. 71013970
Stand and mounting material for one sensor
• for second measuring position; stainless steel 1.4301 (AISI 304)
• order no. 71013968
Flow through assembly
• for bypass applications with open drain
• material: stainless steel 1.4571 (AISI 316 Ti)
• order no. 71013995
Sample reservoir chamber
• dimensions: 540 x 500 x 300 mm (21.3" x 19.7" x 11.8")
• for 1 or 2 sensors
• material: stainless steel 1.4571 (AISI 316 Ti)
• order no. 71013929
Weather protection roof for aluminum version of CAM74
• mandatory for outside use
• dimensions: 370 x 470 x 455 mm (14.6" x 18.5" x 17.9")
• material: polycarbonat (PC)
• order no. 71092182
66
Endress+Hauser
STIP-scan CAM74/CAS74
Troubleshooting
9
Troubleshooting
9.1
Warnings
The following warnings (shown in yellow) may occur without disrupting measurement:
Endress+Hauser
Warning
Possible cause
Corrective measures
NO SLUDGE SETTLING
The sludge is not settling. For this
reason, it is not possible to determine
any further sludge parameters.
Increase the sludge settling time
(SETTINGS menu).
ABSORPTION TOO HIGH
Absorption (average value within the
wavelength range of evaluation)
following the settling process is > 250
m-1.
none
NITRATE VALUE TOO HIGH
The nitrate value is > 23 mg/l. Above
this value, the measuring system
returns values which are too
inaccurate.
none
AIR IN MEASURING CELL
Air has entered the measuring chamber
(e.g. as a result of a drop in the water
level).
If the air has not escaped following four
measurement attempts, the warning is
replaced by an error message, and
operation is suspended.
Lower the sensor a little deeper
into the wastewater so that air
can no longer enter the
measuring cell.
67
Troubleshooting
STIP-scan CAM74/CAS74
9.2
Error messages
Operation is suspended in the case of the following error messages (shown in red):
68
Error message
Possible cause
Corrective measures
NETWORK FAILURE
The transmitter cannot establish
communication with the sensor.
The connection between the control
line and the sensor or the
transmitter is not correct.
Check the communication interface,
the RS232/RS485 converter (LED
must be lit) and the control line.
Are all cables connected?
Contact E+H Service if necessary.
NO COM PORT AVAILABLE
When starting the STIP-scan
software program, no COM
interface was identified.
Check to see which interface the
sensor cable is connected to. Does
this match the entry in the
SETTINGS > INTERFACE menu?
Contact E+H Service if necessary.
LEAKAGE
The leakage sensor signals that there
is water in the sensor.
Contact E+H Service.
Carefully remove the sensor from the
water and store it in an upright
position until a service representative
arrives.
NO LIGHT SIGNAL
If there is a fault in the lamp, the
high-voltage generator or the
spectrometer, there can be no
measuring signal.
Contact E+H Service.
OUTPUT FAILURE
There is a fault in the analog outputs
or the relay contacts.
Contact E+H Service.
STEPPER FAILURE
There is a fault in the motor (e.g.
piston seal is damaged).
Contact E+H Service.
NO VALID REFERENCE
There was a failure in the reference
measurement using distilled water.
Contact E+H Service.
AIR IN MEASURING CELL
Air has entered the measuring
chamber (e.g. as a result of a drop in
the water level).
Lower the sensor a little deeper into
the wastewater.
Contact E+H Service if necessary.
Endress+Hauser
STIP-scan CAM74/CAS74
Troubleshooting
9.3
Spare parts
Description and contents
Order number
Spare parts kit
Protective flange; 3 socket head cap screws M4x8; molded seal 18 x 60 x 0.5 EPDM
71061363
Flow assembly; 3 socket head cap screws M4x8; molded seal 18 x 60 x 0.5 EPDM
71061365
Connecting clip with sealing ring
71061366
9.4
Return
If a repair is needed, please clean the sensor and return it to your sales office.
When returning the item, please use the original packaging.
Please enclose the completed "Declaration of hazardous material and de-contamination" (copy the
second last page of these Operating Instructions) in the packaging together with the dispatch
documents. No repairs can be carried out without a completed declaration!
9.5
Disposal
The device contains electronic components and must therefore be disposed of in accordance with
regulations on the disposal of electronic waste.
Please observe local regulations.
Endress+Hauser
69
Technical Data
STIP-scan CAM74/CAS74
10
Technical Data
10.1
Input
Measuring range
NOx-N
COD equivalent
BOD equivalent
TOC equivalent
SAC254
Sludge TS
SV
SI
ATU
Wavelength
200 to 680 nm
Cable specification
max. 20 m (65.6 ft.)
0.3 to 23 mg/l
10 to 2000 mg/l1)
10 to 2000 mg/l1)
4 to 800 mg/l1)
1 to 250 m-1
0.5 to. 5.0 g/l
100 to 900 ml/l2)
corresponds to SV divided by sludge TS
1 to 200 m-1
1) based on KHP (kalium hydrogen phthalate)
2) undiluted sample
10.2
Output signal
0/4 to 20 mA
Accuracy
±0.1 % of end of measuring range
Burden
max. 500 Ω
Resolution
±0.02 % of end of measuring range
Insulating strength
max. 3000 V DC
Relay
Quantity
Switching current
7 NO contacts
0.5 A at 120 V AC / 1.0 A at 24 V DC
Converter
Transmission rate
Insulating strength
115200 Bit/s
max. 3000 V DC
PROFIBUS
Type
Protocol
I/O memory
PROFIBUS DP slave
DP-V0 or DP-V1 (class 1/2)
368 byte
10.3
70
Output
Power supply
Supply voltage
115/230 V 50/60 Hz (stainless steel version)
100 to 250 V 50 to 60 Hz (aluminum version)
Power consumption
approx. 130 VA
Endress+Hauser
STIP-scan CAM74/CAS74
Technical Data
10.4
Performance characteristics
Measured value
resolution
NOx-N
COD equivalent
BOD equivalent
TOC equivalent
SAC254
0.1 mg/l
2 mg/l
2 mg/l
1 mg/l
0.1 m-1
Measuring interval
Inlet and outlet
Activated sludge basin
2 to 60 min
depending on sludge composition
Repeatability
max. 3 % of the full-scale value for the parameters NOx-N, COD, BOD, TOC, SAC
Response time
120 s (depending on the application; larger intervals can be selected)
10.5
Environment
Ambient temperature
range
-10 to +40 °C (+14 to 104 °F) (transmitter in enclosed weatherproof housing)
Degree of protection
Sensor (when housing is closed)
Transmitter as stainless steel version (closed at the front)
Transmitter as stainless steel version (open at the front)
Transmitter as aluminum version
10.6
Process
Process temperature
range
0 to 30 °C (32 to 86 °F)
If the temperature of the medium > 30 °C, a separate application test is necessary!
Process pressure range
0 to 0.3 bar (4.4 psi) overpressure
Particle size
< 1 mm
Solids content
< 5 g/l (< 5000 ppm)
Sludge volume SV
max. 750 ml/l for undiluted sample after 30 min
10.7
Dimensions
Mechanical construction
Transmitter (stainless steel version)
Transmitter (aluminum version)
Sensor
Endress+Hauser
IP67
IP65
IP55
IP65
W x H x D: 520 x 520 x 260 mm
(20.5" x 20.5" x 10.2")
W x H x D: 327 x 273 x 180 mm
(12.87" x 10.75" x 7.1")
L = approx. 600 mm (23.6"); Ø = 129 mm
(5.08")
Weight
Transmitter (stainless steel version)
Transmitter (aluminum version)
Sensor
approx. 31 kg (68 lbs)
approx. 7.7 kg (17 lbs)
approx. 8.3 kg (18.3 lbs)
Material
Transmitter (stainless steel version)
Transmitter (aluminum version)
Sensor (body)
Sensor holder
stainless steel 1.4301 (AISI 304)
aluminum casting
stainless steel 1.4571 (AISI 316 Ti)
stainless steel 1.4571 (AISI 316 Ti)
71
STIP-scan CAM74/CAS74
Index
A
Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Additional outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Air pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Analog outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
L
66
51
55
46
C
Calibration constants . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Check
Electrical connection. . . . . . . . . . . . . . . . . . . . . . . . . . 24
Installation and function . . . . . . . . . . . . . . . . . . . . . . . 59
Checking
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Cleaning
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 59
Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Configuration of the measuring point . . . . . . . . . . . . . . . . 44
Connection
Transmitter (aluminum version) . . . . . . . . . . . . . . . . . 20
Transmitter (stainless steel version) . . . . . . . . . . . . . . . 16
CSV settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Current parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Current trace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
D
Declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
E
Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Electrical icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Enter measuring function . . . . . . . . . . . . . . . . . . . . . . . . . 39
Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
F
Flow assembly
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
I
Icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Immunity to interference . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 9–11
Flow assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Upright post . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Wall bracket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Leakage test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
M
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Complete measuring point. . . . . . . . . . . . . . . . . . . . . .
Leakage test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measuring point . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measure contamination rate . . . . . . . . . . . . . . . . . . . . . . .
Measured values. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Menu
Additional outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Air pressure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Analog outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Calibration constants . . . . . . . . . . . . . . . . . . . . . . . . . .
Configuration of the measuring point. . . . . . . . . . . . . .
CSV settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Current parameters . . . . . . . . . . . . . . . . . . . . . . . . . . .
Current trace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Enter measuring function . . . . . . . . . . . . . . . . . . . . . .
Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
KHP VALIDATION . . . . . . . . . . . . . . . . . . . . . . . . . . .
Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measure contamination rate . . . . . . . . . . . . . . . . . . . .
Measured values . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Monitoring site . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Naming the measuring point . . . . . . . . . . . . . . . . . . . .
NOx validation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Relay contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Show historical traces . . . . . . . . . . . . . . . . . . . . . . . . .
Single measurement . . . . . . . . . . . . . . . . . . . . . . . . . .
Sludge settling curve . . . . . . . . . . . . . . . . . . . . . . . . . .
Smoothing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Spectrometer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Stepper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Take sample spectra . . . . . . . . . . . . . . . . . . . . . . . . . .
Target/actual adjustment . . . . . . . . . . . . . . . . . . . . . .
Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Test sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Trace pollution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TRACES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Menu
Start measurement . . . . . . . . . . . . . . . . . . . . . . . . . . .
Monitoring site . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
64
64
65
64
56
36
51
55
46
38
42
44
53
27
29
39
50
40
58
56
36
26
28
43
40
49
47
43
33
32
36
45
54
55
41
39
54
57
37
29
26
28
N
Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Naming the measuring point . . . . . . . . . . . . . . . . . . . . . . . 43
NOx validation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
K
KHP VALIDATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
72
Endress+Hauser
STIP-scan CAM74/CAS74
O
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 25–26
Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
P
Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Post-connection check . . . . . . . . . . . . . . . . . . . . . . . . . . .
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
PROFIBUS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
49
24
70
60
Q
Quick commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Quick setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
R
Relay contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5, 69
S
Safety icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Sensor
Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Show historical traces . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Single measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Sludge settling curve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Smoothing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Spectrometer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Start measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Stepper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
T
Take sample spectra . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Target/actual adjustment . . . . . . . . . . . . . . . . . . . . . . . . . 39
Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Test sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Trace pollution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
U
Upright post . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
W
Wall bracket
Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Endress+Hauser
73
STIP-scan CAM74/CAS74
74
Endress+Hauser
a
0
0
0
2
5
8
7
Declaration of Hazardous Material and De-Contamination
Erklärung zur Kontamination und Reinigung
Please reference the Return Authorization Number (RA#), obtained from Endress+Hauser, on all paperwork and mark the RA
clearly on the outside of the box. If this procedure is not followed, it may result in the refusal of the package at our facility.
Bitte geben Sie die von E+H mitgeteilte Rücklieferungsnummer (RA#) auf allen Lieferpapieren an und vermerken Sie diese
auch außen auf der Verpackung. Nichtbeachtung dieser Anweisung führt zur Ablehnung ihrer Lieferung.
RA No.
Because of legal regulations and for the safety of our employees and operating equipment, we need the "Declaration of Hazardous Material
and De-Contamination", with your signature, before your order can be handled. Please make absolutely sure to attach it to the outside of the
packaging.
Aufgrund der gesetzlichen Vorschriften und zum Schutz unserer Mitarbeiter und Betriebseinrichtungen, benötigen wir die unterschriebene
"Erklärung zur Kontamination und Reinigung", bevor Ihr Auftrag bearbeitet werden kann. Bringen Sie diese unbedingt außen an der
Verpackung an.
Type of instrument / sensor
Geräte-/Sensortyp
____________________________________________
Serial number
Seriennummer ________________________
Used as SIL device in a Safety Instrumented System / Einsatz als SIL Gerät in Schutzeinrichtungen
Process data/Prozessdaten
Temperature / Temperatur_____ [°F] _____ [°C]
Conductivity / Leitfähigkeit ________ [µS/cm]
Pressure / Druck
_____ [psi] _____ [ Pa ]
Viscosity /Viskosität _____ [cp] _____ [mm2/s
Medium and warnings
Warnhinweise zum Medium
Medium /concentration Identification flammable
CAS No.
Medium /Konzentration
entzündlich
toxic
giftig
corrosive
ätzend
harmful/
irritant
gesundheitsschädlich/
reizend
other *
harmless
sonstiges* unbedenklic
Process
medium
Medium im
Prozess
Medium for
process cleaning
Medium zur
Prozessreinigung
Returned part
cleaned with
Medium zur
Endreinigung
* explosive; oxidising; dangerous for the environment; biological risk; radioactive
* explosiv; brandfördernd; umweltgefährlich; biogefährlich; radioaktiv
Please tick should one of the above be applicable, include safety data sheet and, if necessary, special handling instructions.
Zutreffendes ankreuzen; trifft einer der Warnhinweise zu, Sicherheitsdatenblatt und ggf. spezielle Handhabungsvorschriften beilegen.
Description of failure / Fehlerbeschreibung __________________________________________________________________________
______________________________________________________________________________________________________________
______________________________________________________________________________________________________________
Company data /Angaben zum Absender
IV
Company /Firma ___________________________________
_________________________________________________
Address / Adresse
_________________________________________________
_________________________________________________
Phone number of contact person /Telefon-Nr. Ansprechpartner:
____________________________________________
Fax / E-Mail ____________________________________________
Your order No. / Ihre Auftragsnr. ____________________________
“We hereby certify that this declaration is filled out truthfully and completely to the best of our knowledge.We further certify that the returned
parts have been carefully cleaned. To the best of our knowledge they are free of any residues in dangerous quantities.”
“Wir bestätigen, die vorliegende Erklärung nach unserem besten Wissen wahrheitsgetreu und vollständig ausgefüllt zu haben. Wir bestätigen
weiter, dass die zurückgesandten Teile sorgfältig gereinigt wurden und nach unserem besten Wissen frei von Rückständen in gefahrbringender Menge sind.”
www.endress.com/worldwide
BA423C/07/en/05.09
71087097
Printed in Germany / FM+SGML 6.0 / DT
71087097