Buhler Technologies Installation- and Operation Instruction Oxygen Flue Gas Analyzer BA 2000
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
Lesen Sie die Bedienungsanleitung vor dem Gebrauch des Gerätes gründlich durch, insbesondere die
Hinweise unter Gliederungspunkt 2. Andernfalls könnten Gesundheits- oder Sachschäden auftreten.
Die Bühler Technologies GmbH haftet nicht bei eigenmächtigen Änderungen des Gerätes oder für
unsachgemäßen Gebrauch.
Read this instruction carefully prior to installation and/or use. Pay attention particularly to all advises
and safety instructions to prevent injuries. Bühler Technologies GmbH can not be held responsible for
misusing the product or unreliable function due to unauthorised modifications.
BX550012, 12/2011
Art. Nr. 90 31 112
Bühler Technologies GmbH, Harkortstr. 29, D-40880 Ratingen
Tel. +49 (0) 21 02 / 49 89-0, Fax. +49 (0) 21 02 / 49 89-20
Internet: www.buehler-technologies.com
Email: analyse@buehler-technologies.com
1
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
Inhaltsverzeichnis
1
Seite
Einleitung ............................................................................................................................. 4
2
Wichtige Hinweise ............................................................................................................... 4
2.1
Allgemeine Gefahrenhinweise ..........................................................................................................5
3
Erläuterungen zum Typenschild ........................................................................................ 6
4
Produktbeschreibung ......................................................................................................... 6
4.1
Allgemeine Beschreibung .................................................................................................................6
4.2
Lieferumfang .....................................................................................................................................6
5
Transport und Lagerungsvorschriften .............................................................................. 7
6
Aufbauen und Anschließen ................................................................................................ 7
6.1
Montage.............................................................................................................................................7
6.2
Anschluss der Gasleitungen .............................................................................................................8
6.2.1
Anschluss der Instrumentenluft für den Injektor ...........................................................................8
6.2.2
Erforderlicher Druck der Instrumentenluft .....................................................................................8
6.2.3
Betrieb des Injektors mit einer Pumpe ..........................................................................................8
6.2.4
Anschluss des Prüfgases ..............................................................................................................8
6.2.5
Anschluss gesonderte Messgasrückführung ................................................................................9
6.3
Elektrische Anschlüsse .....................................................................................................................9
6.3.1
Sondenteil .....................................................................................................................................9
6.3.2
ZrO2-Sensor ..................................................................................................................................9
6.3.3
Netzversorgung und Ausgänge ....................................................................................................9
7
7.1
7.2
7.3
Kalibrierung ....................................................................................................................... 10
Einführung in die Kalibrierung .........................................................................................................10
1-Punkt-Kalibrierung........................................................................................................................10
2-Punkt-Kalibrierung........................................................................................................................11
8
Betrieb ................................................................................................................................ 11
8.1
Funktion des ZrO2-Sensors .............................................................................................................11
8.2
Bedienung des BA 2000-Reglers ....................................................................................................12
8.2.1
Funktion der Tasten ....................................................................................................................12
8.2.2
Übersicht über die Menüführung.................................................................................................13
8.2.3
Ausführliche Erklärung des Bedienungsprinzips ........................................................................14
8.2.4
Beschreibung der Menüfunktionen .............................................................................................15
9
Wartung .............................................................................................................................. 17
9.1
Allgemeine Warnhinweise ...............................................................................................................17
9.2
Auswechseln der Feinsicherung .....................................................................................................18
9.3
Wartung des Filterelementes ..........................................................................................................18
9.3.1
Austrittsfilter.................................................................................................................................18
9.3.2
Eintrittsfilter..................................................................................................................................18
10 Fehlersuche und Beseitigung .......................................................................................... 19
11 Instandsetzung und Entsorgung ..................................................................................... 20
11.1 Entsorgung ......................................................................................................................................20
11.2 Ersatzteil- und Verbrauchsmaterialliste ..........................................................................................20
12 Beigefügte Dokumente ..................................................................................................... 21
13 Anhang ............................................................................................................................... 39
13.1 Belegungsplan.................................................................................................................................39
2
BX550012, 12/2011
Art. Nr. 90 31 112
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
Contents
1
Page
Introduction........................................................................................................................ 22
2
Important Advice ............................................................................................................... 22
2.1
General Indication of Risk ...............................................................................................................23
3
Explanation of the Type Plate .......................................................................................... 24
4
4.1
4.2
Product Description .......................................................................................................... 24
General Description.........................................................................................................................24
Included Items in Delivery ...............................................................................................................24
5
Transport and storing Regulations .................................................................................. 25
6
Installation and Connection ............................................................................................. 25
6.1
Mounting ..........................................................................................................................................25
6.2
Tubing..............................................................................................................................................26
6.2.1
Connection of Instrument Air for the Aspirator ............................................................................26
6.2.2
Required Pressure for Instrument Air .........................................................................................26
6.2.3
Operating the Aspirator with a Pump ..........................................................................................26
6.2.4
Connection of Test Gas ..............................................................................................................26
6.2.5
Connection of separate sample gas recovery ............................................................................27
6.3
Electrical connections .....................................................................................................................27
6.3.1
Probe Section ..............................................................................................................................27
6.3.2
ZrO2-Sensor ................................................................................................................................27
6.3.3
Mains Connection and Outputs ...................................................................................................27
7
Calibration .......................................................................................................................... 28
7.1
Introduction to Calibration ...............................................................................................................28
7.2
1-Point-Calibration...........................................................................................................................28
7.3
2-Point-Calibration...........................................................................................................................29
8
Operation............................................................................................................................ 29
8.1
Function of the ZrO2-Sensor ...........................................................................................................29
8.2
Operation for the BA 2000 Regulator ..............................................................................................30
8.2.1
Key functions ...............................................................................................................................30
8.2.2
Menu of the Regulator.................................................................................................................31
8.2.3
Detailed Description of functional Principle.................................................................................32
8.2.4
Description of the Menu Functions .............................................................................................32
9
Maintenance ....................................................................................................................... 34
9.1
Indication of Risk .............................................................................................................................34
9.2
Replacing the Fuse .........................................................................................................................35
9.3
Maintenance of the Filter Element ..................................................................................................35
9.3.1
Downstream filter ........................................................................................................................35
9.3.2
In-Situ Filter .................................................................................................................................36
10 Troubleshooting and Elimination .................................................................................... 36
11 Repair and Disposal .......................................................................................................... 37
11.1 Disposal ...........................................................................................................................................37
11.2 Spare parts and consumables ........................................................................................................37
12 Attached Documents ......................................................................................................... 38
13 Appendix ............................................................................................................................ 39
13.1 Pin assignment ................................................................................................................................39
BX550012, 12/2011
Art. Nr. 90 31 112
3
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
1 Einleitung
Der Rauchgas-Sauerstoff-Analysator BA 2000 ist zur „quasi-in-situ“ Messung der Sauerstoffkonzentration in
Verbrennungsprozessen bestimmt. Der ZrO2-Sensor ist in einer Sonde montiert, die selbstregelnd auf ca.
180°C beheizt ist, um Kondensation zu vermeiden. Der in das Filtergehäuse integrierte Injektor führt dem
ZrO2-Sensor über ein schnell und einfach zu wechselndes Filterelement kontinuierlich frisches Prozessgas zu
und leitet es anschließend in den Prozess zurück. Der im BA 2000 eingesetzte potentiometrische ZrO2-Sensor
liefert genaue und sehr schnelle Messergebnisse. Für den Betrieb ist kein Referenzgas erforderlich.
Neben der Grundversion BA2000, sind auch die Sonderversionen BA2000-MF
Messgasrückführung) und BA2000-SE (abgesetzte Elektronik bis ca. 15 m) erhältlich.
(gesonderte
Beachten Sie die Zeichnung im Datenblatt und überprüfen Sie vor Einbau des Analysators, ob die genannten
technischen Daten den Anwendungsparametern entsprechen. Überprüfen Sie ebenfalls, ob alle zum
Lieferumfang gehörenden Teile vollständig vorhanden sind.
Bitte beachten Sie beim Anschluss die Kennwerte des Analysators und bei Ersatzteilbestellungen die richtigen
Ausführungen.
2 Wichtige Hinweise
Der Einsatz des Analysators ist nur zulässig, wenn:
- das Produkt unter den in der Bedienungs- und Installationsanleitung beschriebenen Bedingungen, dem
Einsatz gemäß Typenschild und für Anwendungen, für die es vorgesehen ist, verwendet wird.
- die in der Zeichnung und der Anleitung angegebenen Grenzwerte eingehalten werden.
- Überwachungsvorrichtungen/ Schutzvorrichtung korrekt angeschlossen sind.
- die Service- und Reparaturarbeiten, die nicht in dieser Anleitung beschrieben sind, von Bühler
Technologies GmbH durchgeführt werden.
- Originalersatzteile verwendet werden.
Diese Bedienungsanleitung ist Teil des Betriebsmittels. Der Hersteller behält sich das Recht vor, die
Leistungs-, die Spezifikations- oder die Auslegungsdaten ohne Vorankündigung zu ändern. Bewahren Sie die
Anleitung für den späteren Gebrauch auf.
Begriffsbestimmungen für Warnhinweise:
Hinweis
Signalwort für wichtige Information zum Produkt auf die im besonderen Maße aufmerksam
gemacht werden soll.
Vorsicht
Signalwort zur Kennzeichnung einer Gefährdung mit geringem Risiko, die zu einem Sachschaden oder leichten bis mittelschweren Körperverletzungen führen kann, wenn sie nicht
vermieden wird.
Warnung
Signalwort zur Kennzeichnung einer Gefährdung mit mittlerem Risiko, die möglicherweise
Tod oder schwere Körperverletzungen zur Folge hat, wenn sie nicht vermieden wird.
Gefahr
Signalwort zur Kennzeichnung einer Gefährdung mit hohem Risiko, die unmittelbar Tod
oder schwere Körperverletzung zur Folge hat, wenn sie nicht vermieden wird.
4
Warnung vor einer
allgemeinen Gefahr
Warnung vor explosionsgefährdeten Bereichen
Netzstecker ziehen
Warnung vor elektrischer
Spannung
Warnung vor heißer
Oberfläche
Atemschutz tragen
Warnung vor dem
Einatmen giftiger Gase
Gesichtsschutz
tragen
Warnung vor ätzenden
Flüssigkeiten
Handschuhe tragen
BX550012, 12/2011
Art. Nr. 90 31 112
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
2.1 Allgemeine Gefahrenhinweise
Das Gerät darf nur von Fachpersonal installiert werden, das mit den Sicherheitsanforderungen und den
Risiken vertraut ist.
Beachten Sie unbedingt die für den Einbauort relevanten Sicherheitsvorschriften und allgemein gültigen
Regeln der Technik. Beugen Sie Störungen vor und vermeiden Sie dadurch Personen- und Sachschäden.
Der für die Anlage Verantwortliche muss sicherstellen, dass:
- Sicherheitshinweise und Betriebsanleitungen verfügbar sind und eingehalten werden,
- Unfallverhütungsvorschriften der Berufsgenossenschaften beachtet werden; in Deutschland:
Allgemeine Vorschriften” (VBG 1) und “Elektrische Anlagen und Betriebsmittel (VBG 4)”,
- die zulässigen Daten und Einsatzbedingungen eingehalten werden,
- Schutzeinrichtungen verwendet werden und vorgeschriebene Wartungsarbeiten durchgeführt werden,
- bei der Entsorgung die gesetzlichen Regelungen beachtet werden.
Wartung, Reparatur:
- Reparaturen an den Betriebsmitteln dürfen nur von Bühler autorisiertem Personal ausgeführt werden.
- Nur Umbau-, Wartungs- oder Montagearbeiten ausführen, die in dieser Bedienungs- und
Installationsanleitung beschrieben sind.
- Nur Original-Ersatzteile verwenden.
Bei Durchführung von Wartungsarbeiten jeglicher Art müssen die relevanten Sicherheits- und
Betriebsbestimmungen beachtet werden.
GEFAHR
Elektrische Spannung
Gefahr eines elektrischen Schlages.
Trennen Sie das Gerät bei allen Wartungsarbeiten vom Netz. Sichern Sie das
Gerät gegen unbeabsichtigtes Wiedereinschalten.
Das Gerät darf nur von instruiertem, fachkundigem Personal geöffnet werden.
GEFAHR
Giftige, ätzende Gase
Messgas kann gesundheitsgefährdend sein.
Sorgen Sie ggf. für eine sichere Ableitung des Gases.
Stellen Sie vor Beginn der Wartungsarbeiten die Gaszufuhr ab und sichern Sie
sie gegen unbeabsichtigtes Aufdrehen.
Schützen Sie sich bei der Wartung vor giftigen / ätzenden Gasen. Tragen Sie
die entsprechende Schutzausrüstung.
BX550012, 12/2011
Art. Nr. 90 31 112
5
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
GEFAHR
Explosionsgefahr bei Verwendung in Explosionsgefährdeten Bereichen
Das Betriebsmittel ist nicht für den Einsatz in explosionsgefährdeten Bereichen
geeignet.
Durch das Gerät dürfen keine zündfähigen oder explosiven Gasgemische
geleitet werden.
3 Erläuterungen zum Typenschild
Hersteller mit Anschrift
Typbezeichnung
Auftrags-Nr., Artikel-Nr.
Elektrische Versorgung
Messbereich
4 Produktbeschreibung
4.1 Allgemeine Beschreibung
Diese Bedienungs- und Installationsanleitung gilt für den Rauchgas-Sauerstoff-Analysator BA 2000.
Der Analysator kann – je nach Bestellung- mit verschiedenem ab Werk angebautem Zubehör ausgeliefert
sein:
- Austrittsfilter (zum Betrieb erforderlich !)
- Entnahmerohr oder Eintrittsfilter
- Verlängerung für Entnahmerohr/Eintrittsfilter
- Adapterflansch
- Pumpe (falls keine Instrumentenluft zur Verfügung steht)
- Digitalanzeige
Dieses Zubehör ist ebenso wie das separat mitgelieferte Zubehör als gesonderte Position im Auftrag
ausgewiesen.
4.2 Lieferumfang
1x Rauchgas-Sauerstoff-Analysator BA 2000
1x Flanschdichtung und Schrauben
1x Bedienungsanleitung
Angebaute und beiliegende Zubehörteile sind als gesonderte Position im Auftrag ausgewiesen.
6
BX550012, 12/2011
Art. Nr. 90 31 112
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
5 Transport und Lagerungsvorschriften
Den Analysator nur in der Originalverpackung oder einem geeigneten Ersatz transportieren.
Bei längerer Nichtbenutzung ist der Analysator gegen Feuchtigkeit und Wärme zu schützen.
Der Analysator muss in einem überdachten, trockenen, vibrations- und staubfreien Raum bei einer
Temperatur von -20°C bis +60°C aufbewahrt werden.
6 Aufbauen und Anschließen
6.1 Montage
Der Analysator ist zur Flanschmontage vorgesehen. Einbauort und Einbaulage bestimmen sich aus den
anwendungsspezifischen Gegebenheiten. Falls möglich, sollte der Einbaustutzen eine leichte Neigung zur
Kanalmitte haben. Der Einbauort sollte wettergeschützt sein.
Ebenfalls muss auf ausreichenden und sicheren Zugang sowohl für die Installation als auch für spätere
Wartungsarbeiten geachtet werden. Beachten Sie hier insbesondere die Ausbaulänge des Entnahmerohres!
Der Austrittsfilter, der Eintrittsfilter (Option) und das Entnahmerohr (Option) liegen der Lieferung lose bei und
müssen vor Inbetriebnahme installiert werden.
HINWEIS
Der Austrittsfilter und der O-Ring für das Griffstück sind
lose beigelegt und müssen vor Inbetriebnahme eingesetzt
werden (Kapitel 7.4).
Betrieb ohne Austrittsfilter nicht zulässig!
Das Entnahmerohr oder der Eintrittsfilter (falls erforderlich mit der passenden Verlängerung) müssen
eingeschraubt werden. Danach wird der Analysator unter Verwendung der beigefügten Dichtungen und
Schrauben am Gegenflansch befestigt.
Die blank liegenden Flanschteile und ggf. der Einbaustutzen sind nach der Montage vollständig zu isolieren,
damit
Kältebrücken
unbedingt
vermieden
werden.
Das
Isoliermaterial
muss
den
Anwendungsvoraussetzungen entsprechen und wetterfest sein.
HINWEIS
!
Falls es applikationsbedingt zu sehr starker Wärmeabstrahlung im Bereich des Analysators
kommt, ist bauseitig eine entsprechende Abschirmung zum Schutz von Sondenteil, Sensor
und Elektronik anzubringen.
BX550012, 12/2011
Art. Nr. 90 31 112
7
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
6.2 Anschluss der Gasleitungen
6.2.1 Anschluss der Instrumentenluft für den Injektor
Unterhalb der Wetterschutzhaube ist eine Rohrverschraubung zum Anschluss eines Rohres, Ø6mm bzw. ¼“,
für die Instrumentenluft vorgesehen.
Bitte beachten Sie, dass gefilterte Luft mindestens nach PNEUROP / ISO Klasse 4 verwendet wird.
Partikel/ m3
Drucktaupunkt
Restölgehalt
Partikelgröße: (1 bis 5) µm
[°C]
[mg/m3]
≤3
≤5
Klasse
bis 1000
4
(keine Partikel ≥ 15µm)
6.2.2 Erforderlicher Druck der Instrumentenluft
Die Ansprechzeit des Analysators ist abhängig vom Druck der verwendeten Instrumentenluft für den Injektor.
Im angehängten Datenblatt sind Diagramme eingefügt, die die t90-Zeiten in Abhängigkeit vom Volumenstrom
und Druck der Instrumentenluft darstellen.
Aus den Diagrammen wird ersichtlich, dass sich mit zunehmendem Druck der Instrumentenluft (respektive
zunehmendem Volumenstrom) die Ansprechzeit t90 des Analysators verkürzt.
Dem Wunsch nach einer schnellen Ansprechzeit stehen jedoch auch einige Nachteile gegenüber:
- Erhöhter Zellenverschleiß des ZrO2 Sensors (Messgas wird über den Sensor gesaugt)
- Steigende Betriebskosten durch hohen Druckluftverbrauch
- Vermischungsgefahr von angesaugtem Messgas und in den Prozess zurückgeführtem Messgas /
Instrumentenluftgemisches
Wir empfehlen daher eine Druck-Einstellwert vom max. 0,5 bar für die Instrumentenluft (bei laufendem Injektor
gemessen).
6.2.3 Betrieb des Injektors mit einer Pumpe
Wird der Injektor mit einer Pumpe anstelle von Instrumentenluft betrieben, gelten die gleichen Anforderungen
an die angesaugte Luft der Pumpe wie für die Instrumentenluft (siehe 6.2.1).
Auch hier empfehlen wir eine Druck-Einstellwert von max. 0,5 bar auf der Druckseite der Pumpe (bei
laufendem Injektor gemessen).
6.2.4 Anschluss des Prüfgases
Seitlich am Flansch ist eine Rohrverschraubung mit Verschlusskappe zum Anschluss eines Rohres, Ø6mm
bzw. ¼, für das Prüfgas vorgesehen. Während des Betriebes muss der Prüfgasanschluss luftdicht
abgeschlossen sein. Dies kann entweder durch die Verschlusskappe geschehen oder –falls eine permanente
Prüfgasleitung angeschlossen ist- über einen luftdicht schließenden Absperrhahn.
WARNUNG
Gasaustritt
Messgas kann gesundheitsschädlich sein.
Prüfen Sie die Leitungen auf Dichtheit.
8
BX550012, 12/2011
Art. Nr. 90 31 112
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
6.2.5 Anschluss gesonderte Messgasrückführung
Die Sonderversion BA2000-MF ist für die gesonderte Messgasrückführung vorbereitet. Der Anschluss der
Messgasrückführung erfolgt unterhalb der Wetterschutzhaube je nach Ausführung an einem Ø12mm bzw. ½“
Rohr. Es ist dringend darauf zu achten, dass sowohl Druck als auch Strömungsgeschwindigkeit bei
Gasentnahmeort und Rückführung identisch sind.
6.3 Elektrische Anschlüsse
VORSICHT
Elektrische Spannung
Falsche Netzspannung kann das Gerät zerstören
Der Anschluss darf nur von geschultem Fachpersonal vorgenommen werden.
Beachten Sie die auf dem Typenschild angegebene Netzspannung. Achten
Sie auf ausreichende Zugentlastung der Anschlusskabel
WARNUNG
Elektrische Spannung
Beschädigung des Gerätes bei Durchführung der Isolationsprüfung
Führen Sie keine Prüfung der Spannungsfestigkeit mit Hochspannung
am Gesamtgerät durch!
Der Analysator ist mit umfangreichen EMV-Schutzmaßnahmen ausgerüstet. Bei einer Prüfung der
Spannungsfestigkeit werden elektronische Filterbauteile beschädigt. Die notwendigen Prüfungen wurden bei
allen zu prüfenden Baugruppen werkseitig durchgeführt (Prüfspannung je nach Bauteil 1 kV bzw. 1,5 kV).
6.3.1 Sondenteil
Der Sondenteil enthält selbstregelnde Heizelemente. Er wird mit zwei Würfelsteckern nach DIN 43650
ausgeliefert. Ein Stecker dient der Stromversorgung, der andere Stecker führt den Alarmausgang. Die
Stromversorgung ist mit der Elektronik bereits verdrahtet. Der Anschluss für den Alarmausgang erfolgt gemäß
Belegungsplan (siehe Anhang, 13.1). Die Stecker sind so konfiguriert, dass sie beim Anschließen nicht
vertauscht werden können. Diese Konfiguration darf aus Sicherheitsgründen nicht verändert werden.
6.3.2 ZrO2-Sensor
Der Sensor wird mit einer 5-poligen M12-Steckverbindung ausgeliefert und ist mit der Elektronik bereits
verdrahtet.
6.3.3 Netzversorgung und Ausgänge
Im Elektronikgehäuse befindet sich eine Klemmleiste zum Anschluss der Netzversorgung, des
Alarmausgangs und des mA-Signals. Der Anschluss erfolgt gemäß Anschlussdiagramm im Anhang mit den
beigefügten Steckerleisten. Hierzu können die Stecker aus ihrer Fassung herausgenommen und nach der
Verdrahtung wieder eingesteckt werden. Die Anschlussbelegung ist auch auf der Platine aufgedruckt.
BX550012, 12/2011
Art. Nr. 90 31 112
9
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
7 Kalibrierung
Der ZrO2-Sensor muss vor der Kalibrierung mindestens 1 Stunde in Betrieb sein.
7.1 Einführung in die Kalibrierung
Veränderungen der äußeren Bedingungen wie Umgebungstemperatur und Luftdruck, sowie Veränderungen
der Messgasbedingungen wie Durchfluss, Druck und Temperatur können das Messergebnis beeinflussen.
Diese Veränderungen des Messverhaltens nennt man Drift.
Um die Drift zu kompensieren, muss der BA 2000 regelmäßig kalibriert werden. Bei einer Kalibrierung wird
das Messverhalten des BA 2000 mit Prüfgas kontrolliert.
Für den Betrieb des BA 2000 ist eine 1-Punkt-Kalibrierung mit Instrumentenluft völlig ausreichend.
Bei Bedarf kann jedoch auch eine 2-Punkt-Kalibrierung mit Instrumentenluft und einem zusätzlichem Prüfgas
durchgeführt werden. Die O2-Konzentration des Prüfgases sollte idealer Weise der des Messgases
entsprechen.
Wann ist eine Kalibrierung notwendig?
Der BA 2000 sollte in folgenden Fällen kalibriert werden:
¾
Nach jeder Inbetriebnahme nach der Einlaufzeit
¾
Während des Betriebes in regelmäßigen Abständen (14-tägig empfohlen)
7.2 1-Punkt-Kalibrierung
Bei der Kalibrierung muss der Injektor des BA 2000 in Betrieb sein. Der Druck der Instrumentenluft für den
Injektor muss bei der Kalibrierung den Bedingungen der späteren Messung entsprechen.
10
¾
Als Prüfgas Instrumentenluft nach PNEUROP / ISO Klasse 4 (siehe Kapitel 6.2) mit einem Vordruck
von ca. 0,1 bar aufgeben.
¾
Mindestens 5 min warten, bis ein stabiler Endwert erreicht ist.
¾
Im Menü Kalibrierpunkt 1 (CAL1) aufrufen. (Zur Bedienung und Menüführung des BA 2000-Reglers
siehe Kapitel 8).
¾
Wert von 20,9 Vol.-% O2 mit j bestätigen.
¾
Menü mit E verlassen. In der Anzeige erscheint die Konzentration 20,9 (Vol.-% O2).
Hinweis Deaktivierung der 2-Punkt-Kalibrierung: Wurde zuvor eine 2-Punkt-Kalibrierung durchgeführt, bezieht sich die Messwertanpassung auch auf den 2. Kalibrierpunkt. In diesem Fall muss eine
„indirekte“ 2-Punkt-Kalibrierung vorgenommen werden. Dazu unter Menüpunkt CAL2 den O2-Wert auf
0,0% stellen und bestätigen. Die Standard-Parameter werden geladen, die 2-Punkt-Kalibrierung ist
deaktiviert.
BX550012, 12/2011
Art. Nr. 90 31 112
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
7.3 2-Punkt-Kalibrierung
Zusätzlich zu den Schritten der 1-Punkt-Kalibrierung wird hierbei eine Kalibrierung mit einem weiteren Prüfgas
durchgeführt, dessen O2-Konzentration idealer Weise dem späteren Messgas entsprechen sollte.
¾
1-Punkt-Kalibrierung mit Instrumentenluft durchführen
¾
Prüfgas mit bekannter O2-Konzentration mit einem Vordruck von ca. 0,1 bar aufgeben.
¾
Mindestens 5 min warten, bis ein stabiler Endwert erreicht ist. Der angezeigte Wert sollte in etwa der
O2-Konzentration des Prüfgases entsprechen.
¾
Im Menü „Kalibrierpunkt 2“ (CAL2) aufrufen. (Zur Bedienung und Menüführung des BA 2000-Reglers
siehe Kapitel 8).
¾
Konzentration des Prüfgases einstellen und mit j übernehmen.
Hinweis: Bei Einstellen von 0,0 Vol.-% O2 werden die Standard-Parameter geladen, was einer 1Punkt-Kalibrierung entspricht.
¾
Menü mit E verlassen. In der Anzeige erscheint die Konzentration des Prüfgases.
8 Betrieb
8.1 Funktion des ZrO2-Sensors
VORSICHT
Heiße Oberfläche
Verbrennungsgefahr
Bei demontiertem Sensor kann das Sensorrohr bis zu 700°C heiß werden.
Die Elektronik des Sensors regelt die Sensor-Heizung auf 700°C. Das mV-Sensorsignal wird in der Elektronik
verarbeitet und als 4-20 mA-Signal linear ausgegeben (0 … 21 Vol.-% O2 entsprechen 4 … 20 mA).
Während der Aufheizphase (ca. 1 h) werden in der Anzeige abwechselnd „init“ und die Temperatur der
Sensorheizung angezeigt. Bei Erreichen der Betriebstemperatur wechselt die Temperatur- zur
Sauerstoffanzeige und wird abwechselnd mit „init“ angezeigt. Ist die Betriebstemperatur noch nicht erreicht,
liefert das Ausgangssignal 0 mA. Erst bei Erreichen der Betriebstemperatur wird der aktuelle mA-Wert
ausgegeben. Nach Beendigung der Aufheizphase wird die Sauerstoffkonzentration dauerhaft angezeigt.
Die Temperaturanzeige dient zur Überwachung der Sensortemperatur und erscheint bei einer Temperaturabweichung von ± 1,5 K.
Ein Alarm wird ausgegeben bei Unter- oder Überschreiten des fest eingestellten Sollwertes für die Beheizung
des Sensors oder des einstellbaren Sollwertes für die O2-Konzentration.
Status
Anzeige
Ausgabe mA-Signal
Aufheizphase,
Betriebstemperatur noch nicht erreicht
Temperatur abwechselnd mit „init“
0 mA
Aufheizphase,
Betriebstemperatur erreicht
O2-Konzentration abwechselnd mit „init“
aktueller mA-Wert
Aufheizphase beendet,
Temperatur im Sollbereich
Sauerstoffkonzentration
aktueller mA-Wert
Aufheizphase beendet,
Temperatur außerhalb des Sollbereichs
Temperatur blinkt
0 mA
Aufheizphase beendet,
O2- Konzentration außerhalb des
Sollbereichs
O2-Konzentration blinkt
aktueller mA-Wert
BX550012, 12/2011
Art. Nr. 90 31 112
11
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
8.2 Bedienung des BA 2000-Reglers
8.2.1 Funktion der Tasten
Tasten: Die Bedienung erfolgt mit nur 3 Tasten, welche folgende Funktionen haben:
Taste
j
Funktionen
¾ Wechsel von der Messwertanzeige ins Hauptmenü
¾ Auswahl des angezeigten Menüpunktes
¾ Annahme eines editierten Wertes oder einer Auswahl
t
¾ Wechsel zum oberen Menüpunkt
¾ Erhöhen der Zahl beim Ändern eines Wertes oder Wechseln der Auswahl
¾ temporärer Wechsel zur alternativen Messwertanzeige (Option)
b
¾ Wechsel zum unteren Menüpunkt
¾ Erniedrigen der Zahl beim Ändern eines Wertes oder Wechseln der Auswahl
¾ temporärer Wechsel zur alternativen Messwertanzeige (Option)
12
BX550012, 12/2011
Art. Nr. 90 31 112
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
8.2.2 Übersicht über die Menüführung
Anzeige der O2-Konzentration, der Temperatur und des Betriebszustandes
aktuelle O2-Konzentration
Anzeige: xxxx
Angezeigt wird die O2-Konzentration in Schritten von 0,1 Vol.-%. Durch Betätigen der Enter Taste j
gelangt man ins Hauptmenü.
Hauptmenü
Untermenü
Einstellbereich
j
j
BA2000
Anzeige: bA20
Kalibrierpunkt 1
Anzeige: CAL1
20,9 Vol.-% O2
j
Anzeige: 20.9%
t b Wert einstellen
j speichert den Wert
warte 4s: kein speichern
Umgebungsluft, fest
eingestellt auf
20,9 Vol% O2
tb
VD Einstellung
00 0003
Kalibrierpunkt 2
Anzeige: CAL2
0 – 15 Vol.-% O2
j
Anzeige: xxxx
der Prüfgaskonzentration im Bereich:
0,1 – 15 Vol% O2;
0,0 = 1-Punkt-Kalibrierung
tb
Alarmhysterese
Anzeige: A Hi
1 – 22 Vol.-% O2
j
Anzeige: xxxx
j
21 – 0 Vol.-% O2
Anzeige: xxxx
Setzen der oberen
Alarmgrenze im Bereich:
1 – 22 Vol% O2
tb
Alarmhysterese
j
tb
j
Anzeige: A Lo
Setzen der unteren
Alarmgrenze im Bereich:
21 – 0 Vol% O2
tb
mA-Skalierung
Anzeige: nA
Bereich für mA-Output
j
Anzeige: x
tb
Exit Untermenü
Anzeige: E
Globale Einstellungen
Anzeige: toP
Celsius/ Fahrenheit
j
Anzeige: C [F]
Auswahl des Bereiches
für die Skalierung der
mA-Ausgabe (4 – 20mA):
1.0 = 0 – 2,5 Vol.-% O2
2.0 = 0 – 5 Vol.-% O2
3.0 = 0 – 10 Vol.-% O2
4.0 = 0 – 15 Vol.-% O2
5.0 = 0 – max. Vol.-% O2
Einstellen der Anzeigeneinheit. Wahlweise
Celsius oder Fahrenheit
tb
Exit Hauptmenü
Anzeige: E
VD 00 0006
Abb. 1: Menü-Führung BA 2000
BX550012, 12/2011
Art. Nr. 90 31 112
13
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
8.2.3 Ausführliche Erklärung des Bedienungsprinzips
Die ausführliche Erklärung führt Sie Schritt für Schritt durch das Menü des BA 2000.
14
¾
Nach Einschalten des Geräts wird für kurze Zeit die implementierte Software-Version angezeigt.
Anschließend geht das Gerät direkt zur Messwertanzeige über.
¾
Durch Drücken der Taste j gelangt man vom Anzeigemodus ins Hauptmenü. (Es ist gewährleistet,
dass die Regelung auch im Menübetrieb weiter läuft.)
¾
Man navigiert mit den Tasten t b gemäß Abb. 1 durch das Hauptmenü.
¾
Bestätigt man einen Hauptmenüeintrag (j), wird das zugehörige Untermenü aufgerufen.
¾
Hier können Betriebsparameter eingestellt werden. Zum Einstellen der Parameter durchläuft man das
Untermenü mit den Tasten t b und bestätigt mit j den einzustellenden Menüpunkt.
¾
Nun können die Werte innerhalb bestimmter Grenzen durch Betätigen der t b Tasten verstellt
werden. Bestätigt man die Einstellung mit j, wird der eingestellte Wert vom System gespeichert. Im
Anschluss gelangt man automatisch zurück ins Untermenü.
¾
Ein Abbruch oder ein manuelles Rückspringen aus den einstellbaren Bereichen ist nicht vorgesehen.
Wird für einige Sekunden keine Taste gedrückt, wechselt das System zurück ins Untermenü. Die
vorher gesetzten Werte werden nur dann übernommen, wenn sie mit j gespeichert wurden.
¾
Ähnlich verhält es sich mit dem Unter- oder auch mit dem Hauptmenü. Falls das Menü nicht mit E
(Exit) verlassen wird, wechselt das System nach einigen Sekunden selbstständig zurück in den
Anzeigemodus. Dabei werden hier allerdings die zuvor gespeicherten Parameter beibehalten und
nicht wieder zurückgesetzt.
¾
Hinweis: Sobald Werte mit der Enter-Taste gespeichert werden, werden diese für die Reglung
übernommen.
¾
Verlassen des Haupt- bzw. Untermenüs erfolgt durch Auswahl des Menüpunktes E (Exit) und
Bestätigen mit j.
BX550012, 12/2011
Art. Nr. 90 31 112
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
8.2.4 Beschreibung der Menüfunktionen
8.2.4.1 Hauptmenü
Analysator BA 2000 (bA20)
Von hier aus gelangt man zu allen relevanten Einstellmöglichkeiten des
Analysator:
Analysators. Im zugehörigen Untermenü können die Kalibrierpunkte und die
Alarmschwellen ausgewählt werden.
Globale Einstellung (ToP Settings)
Auswahl der globalen Temperatureinheit. Wahlweise Grad Celsius (C) oder
Top Settings
Grad Fahrenheit (F)
Hinweis:
Exit Hauptmenü
Exit
Hinweis.
Zu diesem Hauptmenüpunkt gibt es keinen Untermenüpunkt. Die Einheit der
Temperatur wird direkt ausgewählt werden.
Durch Auswählen gelangt man zurück in den Anzeigemodus.
Beim Wechsel vom Hauptmenü in den Anzeigebereich wird kurzzeitig die
Sensortemperatur angezeigt.
BX550012, 12/2011
Art. Nr. 90 31 112
15
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
8.2.4.2 Untermenü Analysator (Anzeige: bA20)
BA 2000 Æ Kalibrierpunkt 1 (CAL1)
Temperatur
Dieser Wert ist fest eingestellt auf 20,9 Vol.-% O2.
Hinweis:
Bei der Kalibrierung wird dieser Wert übernommen.
BA 2000 Æ Kalibrierpunkt 2 (CAL2)
Temperatur
Diese Einstellung setzt den Wert für die Prüfgas-Konzentration.
Dieser Wert ist kann im Bereich von 0,1 – 15 Vol.-% O2 gesetzt werden.
Anzeige 0,0 Vol.-% O2 entspricht 1-Punkt-Kalibrierung.
Hinweis:
Der Standardwert bei Auslieferung beträgt 0,0%, d.h. werkseitig wurde eine
Kalibrierung mit CAL2 nicht durchgeführt, da die Kalibrierung mit CAL1 völlig
ausreicht.
BA 2000 Æ obere Alarmgrenze (Alarm high)
Hier kann der obere Schwellwert für den optischen Alarm sowie für das
Alarm high
Alarmrelais gesetzt werden. Eingestellt wird die Alarmschwelle im Bereich von
1 – 22 Vol.-% O2.
Hinweis:
Der Standardwert bei Auslieferung beträgt 22 (solange nichts anderes
vereinbart ist).
BA 2000 Æ untere Alarmgrenze (Alarm low)
Hier kann der untere Schwellwert für den optischen Alarm sowie für das
Alarm Low
Alarmrelais gesetzt werden. Eingestellt wird die Alarmschwelle im Bereich von
21 – 0 Vol.-% O2.
Hinweis:
Der Standardwert bei Auslieferung beträgt 0 (solange nichts anderes vereinbart
ist).
mA Skalierung Æ Bereich auswählen (nA)
Auswahl des Bereiches für die Sauerstoffkonzentration, der auf den
mA-Skalierung
Ausgabebereich von 4 bis 20 mA skaliert wird.
Hinweis:
Exit Untermenü
Exit
16
3.0 = 0 – 10 Vol.-% O2
5.0 = 0 – max. Vol.-% O2
1.0 = 0 – 2,5 Vol.-% O2
2.0 = 0 – 5 Vol.-% O2
4.0 = 0 – 15 Vol.-% O2
Der Standardwert bei Auslieferung beträgt 5.0 (solange nichts anderes
vereinbart ist).
Durch Auswählen gelangt man zurück ins Hauptmenü.
BX550012, 12/2011
Art. Nr. 90 31 112
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
9 Wartung
9.1 Allgemeine Warnhinweise
Der Analysator darf nicht außerhalb seiner Spezifikationen betrieben werden.
Beachten Sie die Sicherheitshinweise vor der ersten Inbetriebnahme.
Reparaturen an den Betriebsmitteln dürfen nur von Bühler autorisiertem Personal ausgeführt werden.
Führen Sie nur Umbau-, Wartungs- oder Montagearbeiten aus, die in dieser Bedienungs- und
Installationsanleitung beschrieben sind.
- Verwenden Sie nur Original-Ersatzteile.
- Beachten Sie bei der Durchführung von Wartungsarbeiten jeglicher Art die relevanten Sicherheits- und
Betriebsbestimmungen.
-
GEFAHR
Elektrische Spannung
Gefahr eines elektrischen Schlages.
Trennen Sie das Gerät bei allen Wartungsarbeiten vom Netz. Sichern Sie das
Gerät gegen unbeabsichtigtes Wiedereinschalten.
Der Anschluss darf nur von geschultem Fachpersonal vorgenommen werden.
Achten Sie auf die korrekte Spannungsversorgung!
GEFAHR
Giftige, ätzende Gase
Messgas kann gesundheitsgefährdend sein.
Der Sondenteil darf beim Öffnen nicht unter Druck bzw. Spannung stehen.
Bitte sorgen Sie ggf. für eine sichere Ableitung des Gases.
Schützen Sie sich bei der Wartung, insbesondere beim Filterwechsel vor
giftigen / ätzenden Gasen. Tragen Sie die entsprechende Schutzausrüstung.
Spülen Sie die Gasleitungen vor Beginn der Wartungsarbeiten mit Luft.
VORSICHT
Heiße Oberfläche
Verbrennungsgefahr
Im Betrieb kann je nach Betriebsparametern und Typ eine Gehäusetemperatur
von ca. 100°C entstehen.
Lassen Sie das Gerät erst abkühlen, bevor Sie mit den Wartungsarbeiten
beginnen.
VORSICHT
Überdruck
Der Sondenteil steht im Betrieb unter Druck.
Schließen Sie vor dem Öffnen die Gaszufuhr und machen Sie die Anlage
druckfrei. .
BX550012, 12/2011
Art. Nr. 90 31 112
17
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
9.2 Auswechseln der Feinsicherung
¾
Gerät von der Netzspannung trennen.
¾
Deckel des Elektronikgehäuses öffnen und den unteren Teil der Frontplatte entfernen. Die Sicherung
befindet sich auf der Hauptplatine.
¾
Isolationskappe vom Sicherungshalter nehmen.
¾
Sicherung austauschen und Isolationskappe wieder aufsetzen. Beachten Sie die Netzspannung für
die Auswahl der richtigen Sicherung (siehe auch Kapitel 11.2).
¾
Frontpatte montieren, Deckel schließen und Spannungsversorgung wieder herstellen.
9.3 Wartung des Filterelementes
Der Analysator ist mit einem Partikelfilter ausgerüstet, der je nach Schmutzanfall gewechselt werden muss.
Dazu erforderlichenfalls die Spannungszufuhr unterbrechen und den Prozess abschalten.
9.3.1 Austrittsfilter
¾
Wetterschutzhaube entriegeln und aufstellen.
¾
Den Griff am hinteren Ende des Sondenteils unter leichtem Eindrücken um 90° drehen (Griff muss
dann waagerecht stehen) und herausziehen.
¾
Das verschmutzte Filterelement abziehen und die Dichtflächen kontrollieren.
¾
Vor Aufstecken des neuen Filterelementes, die Dichtung am Griffstopfen (Dichtung gehört zum
Lieferumfang des Filterelementes) erneuern.
¾
Den Griff dann mit neuem Filter einführen und unter leichtem Eindrücken um 90° drehen (Griff muss
dann senkrecht stehen).
¾
Bei herausgenommenem Filter kann erforderlichenfalls auch das Entnahmerohr durch Ausblasen
oder mittels eines Reinigungsstabes von innen gereinigt werden.
Achtung: Hintere Filteraufnahme nicht beschädigen!
HINWEIS
Die Keramikfilterelemente sind von ihrer Beschaffenheit sehr zerbrechlich. Daher die
Elemente vorsichtig handhaben und nicht fallen lassen.
Die Filterelemente aus Edelstahl können in einem Ultraschallbad gereinigt und öfters
wiederverwendet werden, in diesem Falle verwenden Sie auf jeden Fall neue Dichtungen an
Filter und Griffstopfen.
Die Wetterschutzhaube lässt sich nur wieder schließen, wenn der Griff vollständig in der
Senkrechten steht. Dazu die Haube durch leichtes Anheben aus der Verriegelungsstütze
lösen und dann herunterklappen. Auf richtiges Einrasten der Haubenverriegelung achten.
9.3.2 Eintrittsfilter
Der Analysator kann zusätzlich mit einem Eintrittsfilter ausgerüstet werden, welcher sich permanent im
Prozessstrom befindet.
Aufgrund der Prozessbedingungen kann es zum allmählichen Zusetzen des Eintrittsfilters kommen. Sollte
dies der Fall sein, muss der Eintrittsfilter ausgewechselt werden.
Hierzu muss der Analysator vollständig ausgebaut und nach Wechsel des Elementes wie in Kapitel 6
beschrieben wieder installiert werden.
18
BX550012, 12/2011
Art. Nr. 90 31 112
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
10 Fehlersuche und Beseitigung
Problem/Störung
mögliche Ursache
Abhilfe
LED leuchtet nicht
−
Kein Netzanschluss
−
Netzanschluss prüfen
Keine Regleranzeige
−
Sicherung defekt
−
Sicherung überprüfen u. ggf.
wechseln (siehe Kap. 9.3)
−
Regler defekt
−
Gerät einschicken (Kap. 11)
−
Temperaturregler defekt
−
Gerät einschicken (Kap. 11)
−
Heizung defekt
−
Gerät einschicken (Kap. 11)
−
Aufheizphase noch nicht
beendet
−
Aufheizphase abwarten
−
Heizung defekt
−
Gerät einschicken (Kap. 11)
−
Keine / falsche
Spannungsversorgung
−
Spannungsversorgung prüfen
Keine Heizleistung Sonde
−
Heizung defekt
−
Gerät einschicken (Kap. 11)
Vom Sensor kein mA-Signal
−
Aufheizphase noch nicht
beendet
−
Aufheizphase abwarten
−
Heizung defekt
−
Gerät einschicken (Kap. 11)
Kondensatbildung
−
Kältebrücken an der
Entnahmestelle
−
Kältebrücken durch Isolierung
beseitigen
Kein Messgasfluss
−
Gasleitungen undicht
−
Dichtigkeitstest
−
Gasleitung verstopft
−
Gasleitungen prüfen
−
Filterelement verstopft
−
Filterelement reinigen bzw.
austauschen, Entnahmerohr
reinigen
Error 13 an der Regleranzeige
−
wenn Temp. Sollwert nach 1 h
nicht erreicht => Heizung
defekt => Heizung wird
automatisch abgeschaltet.
−
Gerät einschicken (Kap. 11)
Error 06 an der Regleranzeige
−
Temperaturistwert war > 5min
unter 300 °C => Heizung
defekt => Heizung wird
automatisch abgeschaltet.
−
Gerät einschicken (Kap. 11)
Sensor heizt nicht
Temperaturalarm Sonde
BX550012, 12/2011
Art. Nr. 90 31 112
19
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
11 Instandsetzung und Entsorgung
Sollte ein Fehler beim Betrieb auftreten, finden Sie unter Gliederungspunkt 8. Hinweise für die Fehlersuche
und Beseitigung.
Sollten Sie weitere Fragen haben, wenden Sie sich bitte an unseren Service
Tel.: +49-(0)2102-498955 oder Ihre zuständige Vertretung.
Ist nach Beseitigung eventueller Störungen und nach Einschalten der Netzspannung die korrekte Funktion
nicht gegeben, muss das Gerät durch den Hersteller überprüft werden. Bitte senden Sie das Gerät zu diesem
Zweck in geeigneter Verpackung an:
Bühler Technologies GmbH
- Reparatur/Service Harkortstraße 29
40880 Ratingen
Deutschland
Bringen Sie zusätzlich die Dekontaminierungserklärung ausgefüllt und unterschrieben an der Verpackung an.
Ansonsten ist eine Bearbeitung Ihres Reparaturauftrages nicht möglich! Das Formular kann per E-Mail
angefordert werden: service@buehler-technologies.com.
11.1 Entsorgung
Bei der Entsorgung sind die gesetzlichen und Normenregelungen im Anwenderland zu beachten
11.2 Ersatzteil- und Verbrauchsmaterialliste
Bei Ersatzteilbestellungen bitten wir Sie, Gerätetyp und Seriennummer anzugeben.
Bauteile für Nachrüstung und Erweiterung finden Sie im angehängten Datenblatt und in unserem Katalog.
Die folgenden Ersatzteile sollten vorgehalten werden:
Ersatzteil
Artikel-Nr.
Feinsicherung
115V / 230V, 800mA
91 10 00 0001
Flanschdichtung DN65PN6
55 200 09 90 14
Flanschdichtung Entnahmerohr
90 090 68
O-Ring-Satz für Filterelement und Sondenteil, Material: Viton
46 222 01 2
ZrO2-Sensor für BA 2000, komplett mit Dichtungen
55 299 99 9
Anschlusskabel M 12 für Sensor BA 2000
55 200 09 90 17
Filterelement aus Keramik, inkl. O-Ringe aus Viton,
Filtereinheit 3 µm
46 222 02 6
Filterelement aus gesintertem Edelstahl, inkl. O-Ringe aus
Viton, Filtereinheit 5 µm
46 222 01 0
20
BX550012, 12/2011
Art. Nr. 90 31 112
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
12 Beigefügte Dokumente
- Konformitätserklärung
- Datenblatt BA 2000
- Dekontaminierungserklärung
KX550010
DD550012
BX550012, 12/2011
Art. Nr. 90 31 112
21
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
1 Introduction
The oxygen flue gas analyzer BA 2000 is designed for measuring oxygen concentration “quasi in-situ” in
combustion processes. To avoid condensate forming the ZrO2 sensor is mounted inside the self-regulating
probe section, heated up to 180°C (356°F). The aspirator, integrated in the filter housing, continuously
provides a fresh sample gas stream to the ZrO2 sensor. Before entering the measuring cell the gas is filtered.
The filter element can be easily replaced without using tools. The accurate and fast responding ZrO2 sensor
does not need any reference gas.
In addition to the standard BA2000 special versions are also available these being the BA2000-MF (separate
sample gas recovery) and the BA2000-SE (separate electronics up to approx.15 m).
Note the drawings in the attached data sheet and check before installation that the technical data fits with your
application parameters. In addition check also the shipment and make sure you have received all parts.
Regard the specific limits of the analyzer. Please only order the spare parts matching the probe type.
2 Important Advice
Please check before installation of the device that the technical data matches the application parameters. Also
check that the delivery is complete.
Operation of the device is only valid if
- the product is used under the conditions described in the installation- and operation instruction, the
intended application according to the type plate and the intended use,
- the performance limits given in the datasheets and in the installation- and operation instruction are obeyed,
- monitoring devices and safety devices are installed properly,
- service and repair is carried out by Bühler Technologies GmbH, unless described in this manual,
- only original spare parts are used.
This manual is part of the equipment. The manufacturer keeps the right to modify specifications without
advanced notice. Keep this manual for later use.
Definitions for warnings:
NOTE
Signal word for important information to the product.
CAUTION
Signal word for a hazardous situation with low risk, resulting in damaged to the device or
the property or minor or medium injuries if not avoided.
WARNING
Signal word for a hazardous situation with medium risk, possibly resulting in severe
injuries or death if not avoided.
DANGER
Signal word for an imminent danger with high risk, resulting in severe injuries or death if
not avoided.
22
Warning against
hazardous situation
Warning against
possible explosive
atmospheres
disconnect from mains
Warning against
electrical voltage
Warning against hot
surface
wear respirator
Warning against
respiration of toxic
gases
wear face protection
Warning against acid
and corrosive
substances
wear gloves
BX550012, 12/2011
Art. Nr. 90 31 112
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
2.1 General Indication of Risk
Installation of the device shall be performed by trained staff only, familiar with the safety requirements and
risks.
Check all relevant safety regulations and technical indications for the specific installation place. Prevent
failures and protect persons against injuries and the device against damage.
The person responsible for the system must secure that:
-
safety and operation instructions are accessible and followed,
local safety regulations and standards are obeyed,
performance data and installation specifications are regarded,
safety devices are installed and recommended maintenance is performed,
national regulations for disposal of electrical equipment are obeyed.
Maintenance and repair
- Repairs on the device must be carried out by Bühler authorized persons only.
- Only perform modifications, maintenance or mounting described in this manual.
- Only use original spare parts.
During maintenance regard all safety regulations and internal operation instructions.
!
DANGER
Electrical voltage
Electrocution hazard.
Disconnect the device from power supply before stating maintenance. Make
sure that the equipment cannot be reconnected to mains unintentionally.
Installation and maintenance must be carried out by trained staff only. Regard
correct mains supply.
!
DANGER
Toxic and corrosive gases
Sample gas can be hazardous.
Take care that the gas is exhausted in a place where no persons are in
danger.
Before maintenance turn off the gas supply and make sure that it cannot be
turned on unintentionally.
Protect yourself during maintenance against toxic / corrosive gases. Use
gloves, respirator and face protector under certain circumstances.
!
DANGER
Explosion hazard if used in hazardous areas
The device is not suitable for operation in hazardous areas with potentially
explosive atmospheres.
Do not expose the device to combustible or explosive gas mixtures.
BX550012, 12/2011
Art. Nr. 90 31 112
23
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
3 Explanation of the Type Plate
Manufacturer and address
Type declaration
Order no., part no.
Electrical ratings
Measuring range
4 Product Description
4.1 General Description
This operating- and installation manual is for the oxygen flue gas analyzer BA 2000.
Depending on the order, the analyzer may be equipped with optional accessories at the factory, e.g.:
- Downstream filter (necessary for operation!)
- Sample tube or in-situ filter
- Extension for sample tube / in-situ filter
- Adapter flange
- Pump (if instrument air is not available)
- Digital display
These accessories as well as separately delivered accessories are listed on the order.
4.2 Included Items in Delivery
1x Oxygen flue gas analyzer BA 2000
1x Flange gasket and screws
1x Installation- and Operating Instruction
Accessories are listed as separate positions on the order.
24
BX550012, 12/2011
Art. Nr. 90 31 112
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
5 Transport and storing Regulations
The analyzer should be only transported in the original case or in appropriate packing.
Protect analyzer against heat and humidity if not used.
The analyzer must be stored in roofed, dry, vibration- und dust free room. Temperature should be between
-20°C and +60°C (-4°F and 140°F).
6 Installation and Connection
6.1 Mounting
The analyzer is provided with a mounting flange. The installation site and the installation position depend on
the application. Whenever possible the extension tube should point slightly downwards. The place of
installation should be weather shielded.
Make sure that the access to the installation site is safe and free, also for maintenance carried out later and
that there is enough space to take out the probe section even with the extension tube attached.
The downstream filter, the in-situ filter (option) and the sample tube (option) are delivered separately. The
must be installed before getting the equipment started.
NOTE
The downstream filter and the O-ring for the handle are
separate items and have to be inserted into the probe
section.
O-ring handle
Operating without downstream filter is not allowed!
The sample tube or the in-situ filter, if necessary with extension, must be fixed. Finally the sample probe
section is put in place by using the delivered gaskets and bolts.
To avoid cold spots, all metal parts and, if necessary, the mounting flange exposed to ambient must be fully
insulated after the probe section has been installed properly. The type of insulation must be suitable for the
particular application and must be weather proof.
!
HINWEIS
If the application produced large heat radiation at the installation site, probe section, sensor
and electronics must be protected with appropriate shielding.
BX550012, 12/2011
Art. Nr. 90 31 112
25
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
6.2 Tubing
6.2.1
Connection of Instrument Air for the Aspirator
The analyzer is connected to instrument air with fittings placed below the weather shielding to a tube with
diameter Ø6mm or ¼“.
Please only use filtered air according to PNEUROP / ISO Class 4 at least.
Class
4
6.2.2
Particles / m3
Particle size: (1 to 5) µm
Pressure dew point
Oil content
[°C] / [°F]
[mg/m3]
≤ 3 / 37
≤5
up to 1000
( no particles ≥ 15µm)
Required Pressure for Instrument Air
The response time of the analyzer depends largely on the instrument air pressure for the aspirator. The
attached data sheet shows a diagram illustrating the correlation of t90-response time and volume and pressure
of supplied instrument air.
As shown, the response time t90 decreases with increasing pressure (or flow, respectively) of instrument air.
However, the desire for short response time brings about some disadvantages:
- Lower life time of the ZrO2 sensor (sample gas is sucked in through the sensor),
- Higher running costs due to higher consumption of pressurized air,
- Hazard of mixing of sucked in sample gas with a mixture of sample gas/instrument air led back into the
process.
Therefore we recommend limiting the pressure for instrument air to a value of max. 0.5 bar (7.25 psi),
measured with aspirator working.
6.2.3
Operating the Aspirator with a Pump
If the aspirator is operated with a pump instead of instrument air, the same requirements are valid for the
sucked in air as for instrument air (see chapter 6.2.1).
Also we recommend limiting the pressure for instrument air to a value of max. 0.5 bar (7.25 psi), measured
with aspirator working.
6.2.4
Connection of Test Gas
At the side of the flange a fitting with a sealing cap for tubes with diameter Ø6mm or ¼“ is provided to connect
the test gas. During operation the test gas input must be sealed air tight. Use the sealing cap or install an airtight shut-off valve, if the test gas tube is connected permanently.
WARNUNG
Gas leakage
Sample gas may be harmful.
Check the tightness of all tubes.
26
BX550012, 12/2011
Art. Nr. 90 31 112
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
6.2.5
Connection of separate sample gas recovery
Special Version BA2000-MF is prepared for separate sample gas recovery. The connection for the sample
gas recovery is below the weather protection shield to a Ø12mm or ½“ pipe, depending on version. Please
ensure that he pressure and the gas flow rate at the sample point and the return stream destination are
identical.
6.3 Electrical connections
CAUTION
Electrical voltage
Wrong mains voltage may damage the device.
The device must be installed by trained staff only. Regard the mains voltage
as given on the type plate. Install the mains connection strain-relieved.
WARNING
Electrical voltage
Damage to the device during insulation testing.
Do not perform insulation tests with high voltage to the device as a whole.
The analyzer is equipped with extensive EMC protection. If insulation tests are carried out the electronic filter
devices will be damaged. All necessary tests have been carried out for all concerned groups of components at
the factory (test voltage 1 kV or 1.5 kV respectively, depending on the device).
6.3.1
Probe Section
The sample probe section is equipped with self-regulating heating elements. It is delivered with two
connectors according to DIN 43650. One connector is assigned for mains connection, the other is connected
to the alarm outputs. Mains connection is already installed to the electronics. The pin assignment for the alarm
outputs is given in the appendix (see chapter 13.1). Reverse polarity protection avoids the connectors being
plugged in the wrong way round. Do not change this configuration.
6.3.2
ZrO2-Sensor
The sensor is connected via a 5-pole M12-connection and is already wired to the electronics at the factory.
6.3.3
Mains Connection and Outputs
Inside the electronic housing a terminal strip provides the connections for mains supply, alarm signal output
and mA-signal output. The pin assignment is given in the appendix. Use the provided terminal connector.
Unplug the connector from the terminal strip, wire the connection and replace it. The pin assignment is also
printed on the circuit board.
BX550012, 12/2011
Art. Nr. 90 31 112
27
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
7 Calibration
Before starting calibration the ZrO2 sensor must warmed-up for at least 1 hour.
7.1 Introduction to Calibration
Changing of environmental conditions like ambient temperature and barometric pressure as well as changes
of the measuring conditions like gas flow, pressure and temperature may influence the measuring result. This
characteristic is called drift.
To compensate drift, the BA 2000 has to be calibrated in regular intervals. During calibration with a test gas,
the measuring characteristics of the BA 2000 are checked.
1-point-calibration using instrument air is absolutely sufficient.
If needed, a 2-point-calibration may be carried out using instrument air and a test gas. In this case, the O2
concentration of the test gas should correspond to the O2 concentration of the sample gas.
When is calibration of BA 2000 required?
Calibration should be carried in the following cases:
¾
After each initial operation after warm-up
¾
During operation in regular intervals (recommended every two weeks)
7.2 1-Point-Calibration
During calibration the aspirator of the BA 2000 must be in operation. The pressure of the instrument air lead
through the aspirator must correspond to the sample conditions.
28
¾
Supply instrument air according to PNEUROP / ISO Class 4 (see chapter 6.2) as test gas with a
primary pressure of approx. 0.1 bar (1.5 psi).
¾
Wait for at least 5 minutes until a stable measuring value is reached.
¾
Select calibration point 1 (CAL1) from the menu. (For operating and selecting menu items oft the
BA 2000 see chapter 8).
¾
Confirm the value of 20.9 Vol.-% O2 with j.
¾
Quit the menu with E. The display shows the concentration 20.9 Vol.-% O2.
Note: Deactivating 2-point-calibration: If a 2-point-calibration has been carried out earlier, the
calibrated value also relates to the second calibration point. In this case the 2-point-calibration is
deactivated by carrying out an “indirect” 2-point-calibration.
Select menu item CAL2, set the O2 value to 0.0% and confirm the value. The standard parameters are
reloaded and 2-point-calibration is deactivated.
BX550012, 12/2011
Art. Nr. 90 31 112
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
7.3 2-Point-Calibration
In addition to the 1-point-calibration a second measuring point is calibrated using a test gas. The O2
concentration of the test gas should correspond to the O2 concentration of the sample gas.
¾
Carry out 1-point-calibration with instrument air.
¾
Supply test gas of known O2 concentration with a primary pressure of approx. 0.1 bar (1.5 psi).
¾
Wait for at least 5 minutes until a stable measuring value is reached. The displayed value should lie
around the concentration of the test gas.
¾
Select menu item “calibration point 2“ (CAL2). (For operating and selecting menu items of the
BA 2000 see chapter 8).
¾
Set the concentration of the test gas and confirm the value with j.
Note: If a value of 0.0 Vol.-% O2 is set, the standard parameters are reloaded, which is identical to 1point-calibration.
¾
Quit the menu with E. The display shows the concentration of the test gas.
8 Operation
8.1 Function of the ZrO2-Sensor
CAUTION
Hot surface
Burning hazard
If the sensor is demounted, the sensor tube may reach a temperature up to
700°C (1290°F).
The sensor controller regulates the sensor heating to a temperature of 700°C (1290°F). The mV-signal output
of the internal electronic is converted to a linear 4-20 mA-signal output (0 … 21 Vol.-% O2 correspond to 4 …
20 mA).
During warm-up (approx. 1 hour) the display shows alternating “init” and the temperature of the sensor
heating. If operation temperature is reached, the oxygen concentration is shown instead of the temperature.
During warm-up, the mA-signal output is 0 mA. The actual mA-value is available not until the operating
temperature of the measuring cell has been reached. After warm-up the oxygen temperature is shown
permanently.
For monitoring the temperature of the sensor is displayed, if a deviation of more than ± 1.5 K occurs.
The alarm is set if the fixed upper and lower limits for heating or O2-concentration are exceeded.
Status
Display
mA-Signal Output
Warm-up,
Operation temperature not jet reached
Temperature alternating with “init“
0 mA
Warm-up,
Operation temperature reached
O2-concentration alternating with “init“
actual mA-value
Warm-up passed,
Temperature within limits
O2-concentration
actual mA-value
Warm-up passed,
Temperature exceeding limits
Temperature blinks
0 mA
Warm-up passed,
O2- concentration exceeds limits
O2-concentration blinks
actual mA-value
BX550012, 12/2011
Art. Nr. 90 31 112
29
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
8.2 Operation for the BA 2000 Regulator
8.2.1
Key functions
Keys:
Operation uses three keys only, providing the following functions:
Key
Function
j
t
b
30
¾
Switch from measurement display to main menu
¾
Selection of the displayed menu item
¾
Accepting the changed value or selection
¾
Switch to upper menu item
¾
Increase of the value or switching of selection
¾
Temporary switching of alternative measurement display (if option is installed)
¾
Switch to lower menu item
¾
Decrease of the value or switching of selection
¾
Temporary switching of alternative measurement display (if option is installed)
BX550012, 12/2011
Art. Nr. 90 31 112
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
8.2.2
Menu of the Regulator
Display of current temperature and operating state
Current Temperature
Display: xxxx
The display shows the O2-concentration with a resolution of 0.1 Vol.-%. Pressing the Enter button j opens
the Main Menu.
Main Menu
Submenu
Setting Range
BA 2000
Calibration point 1
20.9 Vol.-% O2
j
j
Display: bA20
j
Display: CAL1
Display: 20.9%
tb
adjust value
j
store value
wait 4s: no storage
Ambient air, fixed to 20.9
Vol.-% O2
tb
Calibration point 2
0-15 Vol.-% O2
j
Display: CAL2
Display: xxx
Setting of test gas
concentration in a range:
0.1 – 15 Vol.-% O2;
0.0 = 1-point-calibration
tb
upper alarm threshold
1 – 22 Vol.-% O2
j
Display: A Hi
Display: xxx
Setting upper alarm
threshold in the range:
1 – 22 Vol.-% O2
tb
lower alarm threshold
j
tb
Display: A Lo
21 – 0 Vol.-% O2
j
Display: xxx
j
Setting lower alarm
threshold in the range:
21 - 0 Vol.-% O2
tb
mA-Scaling
Scale for mA-output
Display: nA
Display: x
tb
Set scale for mA-output
(4 – 20 mA):
1.0 = 0 – 2.5 Vol.-% O2
2.0 = 0 – 5 Vol.-% O2
3.0 = 0 – 10 Vol.-% O2
4.0 = 0 – 15 Vol.-% O2
5.0 = 0 – max. Vol.-% O2
Exit Submenu
Display: E
Global Settings
Display: toP
Celsius/ Fahrenheit
j
Display: C [F]
Set unit to Celsius or
Fahrenheit.
tb
Exit Main Menu
Display: E
VE 00 0006
Fig. 1: Menu overview BA 2000
BX550012, 12/2011
Art. Nr. 90 31 112
31
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
8.2.3
Detailed Description of functional Principle
This detailed description leads you through the menu of the BA 2000 step by step.
¾
After switching on the device the implemented software version is displayed for short time. Then the
device switches to measurement display.
¾
Pressing the j key switches from display-mode to main menu. (It is guaranteed that the control of
the device continues during setting-mode.)
¾
To navigate through the main menu use the keys t b as shown in Fig. 1.
¾
To accept the menu item press j and the related submenu is displayed.
¾
Now the parameters may be set. To change the parameters scroll the submenu using the keys t b
and confirm the selected item with j.
¾
The values can be changed within their limits using the keys t b. Pressing the key j stores the
set value. Afterwards the device returns to the submenu automatically.
¾
Wait for a few seconds without pressing any key to return to the submenu without saving the values.
¾
The same procedure holds for the sub- and main menu. If you forget to quit the menu the normal way,
the system returns automatically to display mode. In this case the preset values are kept instead of
being reset.
¾
Note: As soon as the values are saved by pressing the enter key, they are accepted for regulation.
¾
Quit the main menu or the submenu by selecting the menu item E (Exit) and pressing j.
8.2.4
Description of the Menu Functions
8.2.4.1
Main Menu
Analyzer BA 2000 (bA20)
This item allows all relevant settings for the analyzer. In the corresponding
Analyzer :
submenu calibration points and alarm thresholds may be selected.
Global settings (top settings)
Selection of global temperature unit, either degree Celsius (C) or degree
Top Settings
Fahrenheit (F).
Note:
Exit main menu
Exit
Note:
32
This menu has no submenu. The temperature unit is directly selected.
Selecting this item returns to the display mode.
When returning from the main menu to normal operation the sensor
temperature is shown for a short time.
BX550012, 12/2011
Art. Nr. 90 31 112
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
8.2.4.2
Submenu Analyzer (Display: bA20)
BA 2000 Æ calibration point 1 (CAL1)
Temperature
This value is fixed to 20.9 Vol.-% O2.
Note:
During calibration this value must be confirmed.
BA 2000 Æ calibration point 2 (CAL2)
Temperature
This item allows setting of the test gas concentration.
Value ranges from 0.1 to 15 Vol.-% O2.
0.0 Vol.-% O2 is displayed or set is identical to 1-point-calibration.
Note:
Default value at delivery is set to 0.0%, which means that 2-point-calibration
(CAL2) has not been carried out at the factory, because calibration with CAL1 is
absolutely sufficient.
BA 2000 Æ upper alarm limit (Alarm high)
This item allows setting of the upper threshold the alarm relays. The alarm
Alarm high
threshold may be set in the range of 1 – 22 Vol.-% O2.
Note:
Default value at delivery is 22 (unless otherwise agreed).
BA 2000 Æ lower alarm limit (Alarm low)
Alarm Low
This item allows setting of the lower threshold the alarm relays. The alarm
threshold may be set in the range of 21 – 0 Vol.-% O2.
Note:
Default value at delivery is 0 (unless otherwise agreed).
mA Scaling Æ Select scale for mA-output (nA)
Select the range of O2 concentration that is scaled to the output range of 4 to
mA-Output
20 mA.
Note:
Exit submenu
Exit
3.0 = 0 – 10 Vol.-% O2
5.0 = 0 – max. Vol.-% O2
1.0 = 0 – 2.5 Vol.-% O2
2.0 = 0 – 5 Vol.-% O2
4.0 = 0 – 15 Vol.-% O2
Default value at delivery is 5.0 (unless otherwise agreed).
Selecting this item and pressing j returns to the main menu.
BX550012, 12/2011
Art. Nr. 90 31 112
33
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
9 Maintenance
9.1 Indication of Risk
Installation and maintenance of the device shall be performed by trained staff only, familiar with the safety
requirements and risks.
-
The device must be operated within its specifications.
All repairs must be carried out by Bühler authorised personnel only.
Only perform modifications, servicing or mounting described in this manual.
Only use original spare parts.
Regard all relevant safety regulations and internal operating instructions during maintenance.
DANGER
Electrical voltage
Electrocution hazard.
Before any manipulation on the device, disconnect the electrical equipment
from mains power supply. Make sure that the electrical equipment cannot be
reconnected during repair or maintenance.
The wiring must be done by trained staff only. Regard the correct mains
voltage.
DANGER
Toxic, corrosive gases
Sample gas may be harmful.
Before maintenance turn off the gas supply and make sure that it cannot be
turned on unintentionally.
Please exhaust sample gas to a safe place.
Protect yourself against toxic / corrosive gas during maintenance. Use gloves,
respirator and face protector under certain circumstances.
CAUTION
Hot surface
Burning hazard
Depending on the process parameters temperature of the cover may rise to
approx. 100°C (°F) during operation.
Before maintenance let the device cool down.
CAUTION
High pressure
Sample gas probe section is under pressure.
Before opening the probe, shut off gas supply and release the process
pressure.
34
BX550012, 12/2011
Art. Nr. 90 31 112
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
9.2 Replacing the Fuse
¾
Disconnect the device from mains.
¾
Remove the cover of the electronic housing and remove the lower part of the front panel. The fuse is
placed on the main board.
¾
Remove the cap from the fuse holder.
¾
Replace the fuse and re-fix the cap. Regard the electrical ratings to select the correct fuse (see also
chapter 11.2)
¾
Remount the front panel, close the cover and reconnect the mains supply.
9.3 Maintenance of the Filter Element
The analyzer is equipped with a particle filter, which has to be changed in regular intervals depending on the
dust load in the process. Before changing the filter element, shut off the process and disconnect the mains
supply.
9.3.1
Downstream filter
¾
Unlock and open the weather shield.
¾
Press and push slightly the handle on the rear of the probe, turn by 90° and pull (the handle must be
now in horizontal position) the handle with the filter out.
¾
Remove the exhausted filter element and check proper conditions of the sealing surfaces.
¾
Before installing the new filter element always replace the O-Ring on the handle plug carefully, put a
new filter element onto it and insert the handle into the probe.
¾
Push to the very end and turn handle 90° spigot (handle is vertical now).
¾
If the filter is removed, clean the extension tube by blowing with plant air or by mechanical means, if
necessary.
Caution: Do not damage rear element spigot!
NOTE
Ceramic filter elements are very brittle by nature. Handle them with care, don’t let them fall.
Filter elements made out of sintered stainless steel can be cleaned in an ultrasonic bath
and be used several times as long as both seals are still in proper conditions.
The weather shield can only be closed when the handle of the filter is positioned vertically.
Pull cover up to the highest point to unlock support, lower it down in position, ensure that the
latch is locked in.
BX550012, 12/2011
Art. Nr. 90 31 112
35
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
9.3.2
In-Situ Filter
As an option, the analyzer may be equipped with an in-situ filter, staying permanently in the process stream.
Depending on the process conditions, the filter may be clogged after some time and has to be replaced.
For this, the analyzer must be dismounted completely and has to be installed as described in chapter 6 after
replacing the filter element.
10 Troubleshooting and Elimination
Problem / Failure
Possible cause
Solution
LED is off
−
No mains supply
−
Check mains supply
No Display
−
Fuse broken
−
Check fuse and replace it (see
chapter 9.3)
−
Controller defective
−
Return device (Chapter 11)
−
Temperature controller
defective
−
Return device (Chapter 11)
−
Heating defective
−
Return device (Chapter 11)
−
Warm-up not jet finished
−
Wait during warm-up
−
Heating defective
−
Return device (Chapter 11)
−
No / wrong mains supply
−
Check mains supply
No heating of the probe
−
Heating defective
−
Return device (Chapter 11)
Sensor gives no mA-signal
output
−
Warm-up not jet finished
−
Wait during warm-up
−
Heating defective
−
Return device (Chapter 11)
Condensate forming
−
Cold spots at in sampling line
−
Insulate cold spots
no sample gas flow
−
Leakage in gas line
−
Check tightness
−
Tubes clogged
−
Check gas tubes
−
Filter element clogged
−
Replace or clean filter
element, clean sample tube
Error 13 at the display
−
If nominal temp. Value not jet
reached after 1 hour =>
heating defective => heating is
shut off automatically
−
Return device (Chapter 11)
Error 06 at the display
−
Actual temperature value falls
below 300°C (572°F) for more
than 5 minutes => heating
defective => heating is shut off
automatically
−
Return device (Chapter 11)
Sensor does not heat up
Temperature alarm probe
36
BX550012, 12/2011
Art. Nr. 90 31 112
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
11 Repair and Disposal
If the device shows irregularities see chapter 8 for troubleshooting
If you need help or more information
call +49(0)2102-498955 or your local agent.
If the device doesn’t work correctly after elimination of failures and turning power on, the device must be
checked by the manufacturer. Please ship the device with suitable packing to
Bühler Technologies GmbH
- Service Harkortstraße 29
40880 Ratingen
Germany
In Addition, attach the filled in and signed Declaration of Decontamination status to the packing. Otherwise,
your repair order cannot be processed! The form can be requested by e-mail to service@buehlertechnologies.com.
11.1 Disposal
Regard the local regulations for disposal of electric and electronic equipment.
11.2 Spare parts and consumables
To order spare parts please indicate type of cooler and serial no. For accessories and enhancement see data
sheets and/or catalogue.
Accessories are listed in the datasheet attached and in our catalogue.
The following parts are recommended for stocking:
Spare part
Part no.
Micro fuse, slow blow
115V / 230V, 800 mA
91 10 00 0001
Flange gasket DN65PN6
55 200 09 90 14
Flange gasket sample tube
90 090 68
O-ring-set for filter element and probe, material: Viton
46 222 01 2
ZrO2-sensor for BA 2000, complete incl. gaskets
55 299 99 9
Connection cable M 12 for sensor BA 2000
55 200 09 90 17
Sintered ceramic filter element, incl. O-rings, material:
Viton, retention rate: 3µm
46 222 02 6
Sintered stainless steel element, incl. O-rings,
material: Viton retention rate 5µm
46 222 01 0
BX550012, 12/2011
Art. Nr. 90 31 112
37
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
12 Attached Documents
- Declaration of conformity
- Data sheet BA 2000
- Declaration of Contamination status
38
BX550012, 12/2011
KX550010
DE550012, DA550012
Art. Nr. 90 31 112
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Rauchgas-Sauerstoff-Analysator / Oxygen Flue Gas Analyzer
BA 2000
13 Anhang
13 Appendix
13.1 Belegungsplan
13.1 Pin assignment
BX550012, 12/2011
Art. Nr. 90 31 112
39
Analysentechnik
Rauchgas-Sauerstoff-Analysator BA 2000
In einigen Verbrennungsprozessen wie z.B. Prozesserhitzer,
Dampfkessel oder Heizöfen kann der Luftbedarf zur Erreichung eines optimalen Wirkungsgrads der Anlage schnell
schwanken. Dabei bewegt sich das wirtschaftliche Optimum
des Prozesses innerhalb eines relativ schmalen Korridors. Sowohl eine erhöhte NOx bzw. SOx Emission, verursacht durch
ein zu großes Luftangebot (O2 Überschuss), als auch ein Energieverlust der Anlage durch eine unvollständige Verbrennung
(O2 Mangel) erfordert eine Sauerstoffmessung im Rauchgas
des Verbrennungsprozesses. Die Probenentnahme nahe an
der Brennkammer ist daher ebenso erforderlich wie ein
schnell ansprechender Sensor, damit auf Änderungen hinsichtlich der Brenngaszusammensetzung und/oder anderer
Variablen des Verbrennungsprozesses unmittelbar reagiert
werden kann.
Der BA 2000 ist speziell für diese Anwendung entwickelt worden.
Schnelle Ansprechzeit
Filterwechsel ohne Werkzeug
Einfache Handhabung
Rauchgastemperaturen bis 1600 °C
Langlebige ZrO2-Messzelle
Display mit O2-Anzeige
4-20 mA Ausgangssignal
Umgebungstemperatur -20 bis +70 °C
Kein Referenzglas erforderlich
Kein Prüfgas erforderlich
Keine Gasaufbereitung erforderlich
Kalibrierung mit Instrumentenluft
DD550012
03/2017
Seite 1 / 6
Bühler Technologies GmbH, Harkortstr. 29, D-40880 Ratingen
Tel. +49 (0) 21 02 / 49 89-0, Fax: +49 (0) 21 02 / 49 89-20
E-Mail: analyse@buehler-technologies.com
Internet: www.buehler-technologies.com
BA 2000
Beschreibung
Der in das Filtergehäuse integrierte Injektor führt dem ZrO2-Sensor kontinuierlich frisches Prozessgas zu. Der Sondenteil wird
selbstregelnd auf 180 °C beheizt, um Kondensation zu vermeiden. Der im BA 2000 eingesetzte, ZrO2-Sensor liefert genaue und
sehr schnelle Messergebnisse.
Für den Betrieb ist kein Referenzgas erforderlich. Die 1-Punkt-Kalibrierung des Sensors erfolgt mit Instrumentenluft, die ebenfalls für den Betrieb des Injektors erforderlich ist. Bei Bedarf kann jedoch auch eine 2-Punkt-Kalibrierung durchgeführt werden.
Das Prüfgas, mit dem hierbei zusätzlich kalibriert wird, sollte idealerweise der O2-Konzentration des Messgases entsprechen.
Die Messzelle wird durch einen im Sondenteil integrierten Filter vor Staubbelastung geschützt. Der Filter ist ohne Werkzeug innerhalb von wenigen Sekunden durch einfache 90°-Drehung des Griffstückes zu wechseln.
Die Filtereinsätze gibt es in Keramik, gesinterter oder sterngefalteter Edelstahl-Ausführung.
Der BA 2000 kann in Zusammenhang mit den aufgeführten Filtern bei Gasen mit einer Staubbeladung bis ca. 2 g/m³ eingesetzt
werden.
Der BA 2000 stellt alle zum sicheren Betrieb benötigten Informationen zur Verfügung. Die Steuerung verfügt über eine Anzeige
mit Tastenfeld zur Befehlseingabe, Alarmausgang, Kalibrierfunktion und 4 - 20 mA Ausgangssignal.
t 90 [sec]
t 90 [sec]
t 90 [sec]
t90-Zeiten in Abhängigkeit vom Volumenstrom und Druck
Volumenstrom Messgas [l/h]
Volumenstrom Instrumentenluft [l/h]
Druck Instrumentenluft [bar]
bei 1 bar absolutem Prozessdruck
Messprinzip der Sauerstoffmessung mit ZrO2-Zellen
Als Grundlage für die Bestimmung der Sauerstoffkonzentration in Gasen mit Zirkoniumdioxidmesszelle dient die NERNSTGleichnung.
(l) U =
RT
4F
In
p (O2 ), Luft
p (O2 ), Messgas
U
R
T
F
p (O2 ), Messgas
p (O2 ), Luft
Zellspannung in V
Molare Gaskonstante, R = 8,31447 J/(molּK)
Messtemperatur in K
Faraday-Konstante, F = 96485,34 C/mol
Partialdruck des Sauerstoffes an der Bezugselektrode in trockener Luft in Pa
Partialdruck des Sauerstoffes an der Bezugselektrode in trockener Luft in Pa
Die Oxidionenleitfähigkeit von Zirkomiumdioxid steigt exponentiell mit der Temperatur an und erreicht oberhalb von 600 °C
genügend große Werte.
Unter der Voraussetzung, dass die Gesamtdrücke der Gase an beiden Elektronen in etwa gleich groß sind (in diesem Fall kann
man mit Volumenkonzentrationen anstelle der Partialdrücke rechnen), ergibt sich nach Einsetzen der Zahlenwerte für die Konstanten in Gleichung (I) folgende Bestimmungsgleichung für die Sauerstoffkonzentration.
(lI)
φ (O2 ) = 20,9
(-46, 42
U
T
)
φ (O2 )
U
T
20,9
Sauerstoffkonzentration im Messgas in Vol.-%
Potentialdifferenz in mV
Messtemperatur in K
Sauerstoffkonzentration in trockener Luft in Vol.-%
Im BA 2000 wird eine Potentiometrische Zelle eingesetzt. Bezugs- und Messgaselektrode befinden sich in zwei unterschiedlichen Gasräumen mit unterschiedlichem Sauerstoffpartialdruck. Die beiden Räume werden durch das gasdichte ZrO2-Rohr getrennt. An den Elektroden entsteht eine EMK (elektromotorische Kraft), die der Partialdruckdifferenz des Sauerstoffs proportional ist. Es gilt die NERNST-Gleichung.
2
Bühler Technologies GmbH
Technische Änderungen vorbehalten.
DD550012 ◦ 03/2017
BA 2000
Technische Daten
Technische Daten BA 2000
Entnahmerohrlänge:
0,5 ... 2 m
Netzspannung:
115 oder 230 V, 50/60 Hz
Heizleistung Sonde:
400 W
Messbereich:
0,1 bis 21 Vol.-% O2
Ausgangssignal:
4-20 mA = 0-21 Vol.-% O2 (skalierbar 0-2,5/0-5/0-10/0-15)
Genauigkeit:
relativer Fehler < 5 %
T90-Zeit Sensor:
< 15 sec
Alarm Sensor:
Unter- oder Überschreiten Heizung Sollwert (fest eingestellt)
Unter- oder Überschreiten O2-Konzentration (einstellbar)
Alarm Sonde:
Untertemperatur
Umgebungstemperatur:
-20 … +70 °C
Prozesstemperatur:
bis zu 1600 °C, je nach Entnahmerohr
Betriebstemperatur Sonde:
max. 200 °C
Material Sonde:
1.4571
Prüfgas 1 Punktkalibration:
Instrumentenluft 20,9 Vol.-% O2
Prüfgase 2 Punktkalibration:
Instrumentenluft 20,9 Vol.-% O2 und
Prüfgas 0,1 bis 15 Vol.-% O2
Bestellhinweise
Artikel-Nr.
Bezeichnung
55200099
BA 2000, 230 V 50/60Hz
55201099
BA 2000-MF, 230 V 50/60Hz
55202099
BA 2000-SE, 230 V 50/60Hz
55200098
BA 2000, 115 V 50/60Hz
55201098
BA 2000-MF, 115 V 50/60Hz
55202098
BA 2000-SE, 115 V 50/60Hz
55200098I
BA 2000I, 115 V 50/60Hz, US sized
55201098I
BA 2000I-MF, 115 V 50/60Hz, US sized
55202098I
BA 2000I-SE, 115 V 50/60Hz, US sized
55200099I
BA 2000I, 230 V 50/60Hz, US sized
55201099I
BA 2000I-MF, 230 V 50/60Hz, US sized
55202099I
BA 2000I-SE, 230 V 50/60Hz, US sized
MF = gesonderte Messgasrückführung
SE = abgesetzte Elektronik bis ca. 15 m
Adapterflansche
Artikel-Nr.
Bezeichnung
55200001
Adapterflansch DN65 PN6 auf Servomex
55200002
Adapterflansch DN65 PN6 auf Thermox
55200001I
Adapterflansch DN3-150 auf Servomex
55200002I
Adapterflansch DN3-150 auf Thermox
DD550012 ◦ 03/2017
Technische Änderungen vorbehalten.
Bühler Technologies GmbH
3
BA 2000
Zeichnungen
4
Bühler Technologies GmbH
Technische Änderungen vorbehalten.
DD550012 ◦ 03/2017
BA 2000
DD550012 ◦ 03/2017
Technische Änderungen vorbehalten.
Bühler Technologies GmbH
5
BA 2000
6
Bühler Technologies GmbH
Technische Änderungen vorbehalten.
DD550012 ◦ 03/2017
Gas Analysis
Oxygen Flue Gas Analyzer BA 2000
Some combustion processes, e.g. process heaters, steam boilers or heating furnaces the air required to achieve optimal
system efficiency can easily fluctuate. And the economic optimum of the process fluctuates within a relatively narrow
range. Both elevated NOx or SOx emission due to an excess
supply of air (excess O2) as well as energy loss within the system due to incomplete combustion (lack of O2) require measuring the oxygen level in the flue gas of the combustion process. Sampling near the combustion chamber is therefore just
as vital as using a rapid response sensor to allow for responding to changes in the combustion gas temperature and/or
other variables in the combustion process promptly.
The BA 2000 was developed specifically for this application.
Quick response time
Tool-less filter change
Easy handling
Flue gas temperatures up to 1600 °C
Durable ZrO2 measuring cell
Display includes O2
4-20 mA output signal
Ambient temperature -20 to +70 °C
No reference glass needed
No test gas needed
No gas conditioning needed
Calibration with instrument air
DE550012
03/2017
page 1 / 6
Bühler Technologies GmbH, Harkortstr. 29, D-40880 Ratingen
Tel. +49 (0) 21 02 / 49 89-0, Fax: +49 (0) 21 02 / 49 89-20
E-Mail: analyse@buehler-technologies.com
Internet: www.buehler-technologies.com
BA 2000
Description
The injector built into the filter housing constantly supplies the ZrO2 sensor with fresh process gas. The self-regulating probe
part is heated to 180 °C to prevent condensation. The ZrO2 sensor in the BA 2000 provides accurate, extremely fast measurements.
No reference gas needed for operation. The sensor's 1-point calibration uses instrument air also needed to operate the injector. If
necessary, 2-point calibration may also be performed. The test gas additionally needed for calibration in this case should ideally
correspond with the O2 concentration of the sample gas.
A filter built into the probe protects the measuring cell from dust exposure. The filter can be changed in seconds without tools
by simply turning the handle 90°.
The filter elements are available in ceramic, sintered or star-pleated stainless steel.
Together with the filters listed, the BA 2000 can be used for gases with a dust load of up to approx. 2 g/m³.
The BA 2000 will provide all the information required for safe operation. The controller features a display with key pad for entering commands, alarm output, calibration function and 4 - 20 mA output signal.
t 90 [sec]
t 90 [sec]
t 90 [sec]
t90 times depending on volume flow and pressure
Sample gas volume flow [L/h]
Instrument air volume flow [L/h]
Instrument air pressure [bar]
at a process pressure of 1 bar absolute
Oxygen measurement principle with ZrO2 cells
The NERNST equation serves as the basis for determining the oxygen concentration in gases by zirconium dioxide measuring
cell.
(L) U =
RT
4F
In
p (O2 ), air
p (O2 ), sample gas
U
R
T
F
p (O2 ), sample gas
p (O2 ), air
Cell voltage in V
Universal gas constant, R = 8.31447 J/(molּK)
Measuring temperature in K
Faraday-constant, F = 96485.34 C/mol
Partial pressure of oxygen at the reference electrode in dry air in Pa
Partial pressure of oxygen at the reference electrode in dry air in Pa
The conductivity of the oxide ions of zirconium oxide increases exponentially with the temperature and reaches adequate values above 600 °C.
Provided the total pressures of the gases are about equal on both electrodes (in this case volume concentrations can be used in
place of partial pressures), after adding the numeric values for the constants in equation (I) results in the following equation for
the oxygen concentration.
(II)
φ (O2 ) = 20.9
(-46, 42
U
T
)
φ (O2 )
U
T
20.9
Oxygen concentration in sample gas in Vol.-%
Potential difference in mV
Measuring temperature in K
Oxygen concentration in dry air in Vol.-%
The BA 2000 uses a potentiometric cell. The reference and the sample gas electrode are located in two different gas chambers
with different oxygen partial pressure. The two chambers are separated by the gas-tight ZrO2 tube. The electrodes generate
e.m.f. (electromotive force) proportional to the partial pressure difference of the oxygen. The NERNST equation applies.
2
Bühler Technologies GmbH
We reserve the right to amend specification.
DE550012 ◦ 03/2017
BA 2000
Technical Data
BA 2000 Technical Data
Sampling tube length:
0.5 ... 2 m
Voltage:
115 or 230 V, 50/60 Hz
Probe heat output:
400 W
Measuring range:
0.1 to 21 Vol.-% O2
Output signal:
4-20 mA = 0-21 Vol.-% O2 (scalable 0-2.5/0-5/0-10/0-15)
Accuracy:
relative error < 5 %
Sensor T90 time:
< 15 sec
Alarm Sensor:
Upper and lower limit of nominal value for heating (fixed)
Upper and lower limit of O2 concentration (adjustable)
Probe alarm:
Insufficient temperature
Ambient temperature:
-20 … +70 °C
Process temperature:
up to 1600 °C, depending on sampling tube
Probe operating temperature:
max. 200 °C
Probe material:
1.4571
Test gas 1-point calibration:
Instrument air 20.9 Vol.-% O2
Test gases 2-point calibration:
Instrument air 20.9 Vol.-% O2 and
test gas 0.1 to 15 Vol.-% O2
Ordering instructions
Item no.
Description
55200099
BA 2000, 230 V 50/60Hz
55201099
BA 2000-MF, 230 V 50/60Hz
55202099
BA 2000-SE, 230 V 50/60Hz
55200098
BA 2000, 115 V 50/60Hz
55201098
BA 2000-MF, 115 V 50/60Hz
55202098
BA 2000-SE, 115 V 50/60Hz
55200098I
BA 2000I, 115 V 50/60Hz, US sized
55201098I
BA 2000I-MF, 115 V 50/60Hz, US sized
55202098I
BA 2000I-SE, 115 V 50/60Hz, US sized
55200099I
BA 2000I, 230 V 50/60Hz, US sized
55201099I
BA 2000I-MF, 230 V 50/60Hz, US sized
55202099I
BA 2000I-SE, 230 V 50/60Hz, US sized
MF = separate sample gas recovery
SE = separated electronics up to approx. 15 m
Adapter flanges
Item no.
Description
55200001
Adapter flange DN65 PN6 to Servomex
55200002
Adapter flange DN65 PN6 to Thermox
55200001I
Adapter flange DN3-150 to Servomex
55200002I
Adapter flange DN3-150 to Thermox
DE550012 ◦ 03/2017
We reserve the right to amend specification.
Bühler Technologies GmbH
3
BA 2000
Drawings
4
Bühler Technologies GmbH
We reserve the right to amend specification.
DE550012 ◦ 03/2017
BA 2000
DE550012 ◦ 03/2017
We reserve the right to amend specification.
Bühler Technologies GmbH
5
BA 2000
6
Bühler Technologies GmbH
We reserve the right to amend specification.
DE550012 ◦ 03/2017
Gas Analysis
Oxygen Flue Gas Analyzer BA 2000
Some combustion processes, e.g. process heaters, steam boilers or heating furnaces the air required to achieve optimal
system efficiency can easily fluctuate. And the economic optimum of the process fluctuates within a relatively narrow
range. Both elevated NOx or SOx emission due to an excess
supply of air (excess O2) as well as energy loss within the system due to incomplete combustion (lack of O2) require measuring the oxygen level in the flue gas of the combustion
process. Sampling near the combustion chamber is therefore
just as vital as using a rapid response sensor to allow for responding to changes in the combustion gas temperature and/
or other variables in the combustion process promptly.
The BA 2000 was developed specifically for this application.
Quick response time
Tool-less filter change
Easy handling
Flue gas temperatures up to 1600 °C (2912 °F)
Durable ZrO2 measuring cell
Display includes O2
4-20 mA output signal
Ambient temperature -20 to +70 °C (-4 to 158 °F)
No reference glass needed
No test gas needed
No gas conditioning needed
Calibration with instrument air
DA550012
03/2017
page 1 / 6
Buhler Technologies LLC, 1030 West Hamlin Road, Rochester Hills, MI 48309
Phone: 248.652.1546 / 49 89-0, Fax: 248.652.1598
e-mail: sales@buhlertech.com
Internet: www.buhlertech.com
BA 2000
Description
The injector built into the filter housing constantly supplies the ZrO2 sensor with fresh process gas. The self-regulating probe
part is heated to 180 °C (356 °F) to prevent condensation. The ZrO2 sensor in the BA 2000 provides accurate, extremely fast measurements.
No reference gas needed for operation. The sensor's 1-point calibration uses instrument air also needed to operate the injector. If
necessary, 2-point calibration may also be performed. The test gas additionally needed for calibration in this case should ideally
correspond with the O2 concentration of the sample gas.
A filter built into the probe protects the measuring cell from dust exposure. The filter can be changed in seconds without tools
by simply turning the handle 90°.
The filter elements are available in ceramic, sintered or star-pleated stainless steel.
Together with the filters listed, the BA 2000 can be used for gases with a dust load of up to approx. 2 g/m³.
The BA 2000 will provide all the information required for safe operation. The controller features a display with key pad for entering commands, alarm output, calibration function and 4 - 20 mA output signal.
t 90 [sec]
t 90 [sec]
t 90 [sec]
t90 times depending on volume flow and pressure
Sample gas volume flow [L/h]
Instrument air volume flow [L/h]
Instrument air pressure [bar]
at a process pressure of 1 bar (14.5 psi) absolute
Oxygen measurement principle with ZrO2 cells
The NERNST equation serves as the basis for determining the oxygen concentration in gases by zirconium dioxide measuring
cell.
(L) U =
RT
4F
In
p (O2 ), air
p (O2 ), sample gas
U
R
T
F
p (O2 ), sample gas
p (O2 ), air
Cell voltage in V
Universal gas constant, R = 8.31447 J/(molּK)
Measuring temperature in K
Faraday-constant, F = 96485.34 C/mol
Partial pressure of oxygen at the reference electrode in dry air in Pa
Partial pressure of oxygen at the reference electrode in dry air in Pa
The conductivity of the oxide ions of zirconium oxide increases exponentially with the temperature and reaches adequate values above 600 °C (1112 °F).
Provided the total pressures of the gases are about equal on both electrodes (in this case volume concentrations can be used in
place of partial pressures), after adding the numeric values for the constants in equation (I) results in the following equation for
the oxygen concentration.
(II)
φ (O2 ) = 20.9
(-46, 42
U
T
)
φ (O2 )
U
T
20.9
Oxygen concentration in sample gas in Vol.-%
Potential difference in mV
Measuring temperature in K
Oxygen concentration in dry air in Vol.-%
The BA 2000 uses a potentiometric cell. The reference and the sample gas electrode are located in two different gas chambers
with different oxygen partial pressure. The two chambers are separated by the gas-tight ZrO2 tube. The electrodes generate
e.m.f. (electromotive force) proportional to the partial pressure difference of the oxygen. The NERNST equation applies.
2
Buhler Technologies LLC
We reserve the right to amend specification.
DA550012 ◦ 03/2017
BA 2000
Technical Data
BA 2000 Technical Data
Sampling tube length:
0.5...2 m (1.6...6.6 ft)
Voltage:
115 or 230 V, 50/60 Hz
Probe heat output:
400 W
Measuring range:
0.1 to 21 Vol.-% O2
Output signal:
4-20 mA = 0-21 Vol.-% O2 (scalable 0-2.5/0-5/0-10/0-15)
Accuracy:
relative error < 5 %
Sensor T90 time:
< 15 sec
Alarm Sensor:
Upper and lower limit of nominal value for heating (fixed)
Upper and lower limit of O2 concentration (adjustable)
Probe alarm:
Insufficient temperature
Ambient temperature:
-20 … +70 °C (-4 … 158 °F)
Process temperature:
up to 1600 °C (2912 °F), depending on sampling tube
Probe operating temperature:
max. 200 °C (392 °F)
Probe material:
1.4571
Test gas 1-point calibration:
Instrument air 20.9 Vol.-% O2
Test gases 2-point calibration:
Instrument air 20.9 Vol.-% O2 and
test gas 0.1 to 15 Vol.-% O2
Ordering instructions
Item no.
Description
55200099
BA 2000, 230 V 50/60Hz
55201099
BA 2000-MF, 230 V 50/60Hz
55202099
BA 2000-SE, 230 V 50/60Hz
55200098
BA 2000, 115 V 50/60Hz
55201098
BA 2000-MF, 115 V 50/60Hz
55202098
BA 2000-SE, 115 V 50/60Hz
55200098I
BA 2000I, 115 V 50/60Hz, US sized
55201098I
BA 2000I-MF, 115 V 50/60Hz, US sized
55202098I
BA 2000I-SE, 115 V 50/60Hz, US sized
55200099I
BA 2000I, 230 V 50/60Hz, US sized
55201099I
BA 2000I-MF, 230 V 50/60Hz, US sized
55202099I
BA 2000I-SE, 230 V 50/60Hz, US sized
MF = separate sample gas recovery
SE = separated electronics up to approx. 15 m (49.2 ft)
Adapter flanges
Item no.
Description
55200001
Adapter flange DN65 PN6 to Servomex
55200002
Adapter flange DN65 PN6 to Thermox
55200001I
Adapter flange DN3-150 to Servomex
55200002I
Adapter flange DN3-150 to Thermox
DA550012 ◦ 03/2017
We reserve the right to amend specification.
Buhler Technologies LLC
3
BA 2000
Drawings
4
Buhler Technologies LLC
We reserve the right to amend specification.
DA550012 ◦ 03/2017
BA 2000
DA550012 ◦ 03/2017
We reserve the right to amend specification.
Buhler Technologies LLC
5
BA 2000
6
Buhler Technologies LLC
We reserve the right to amend specification.
DA550012 ◦ 03/2017
RMA - Dekontaminierungserklärung
RMA - Decontamination Statement
DE/EN
Gültig ab / valid since: 2014/11/01
Revision / Revision 1
ersetzt Rev. / replaces Rev. 0
Um eine schnelle und reibungslose Bearbeitung Ihres Anliegens zu erreichen, füllen Sie bitte diesen Rücksendeschein
aus. Eine genaue Fehlerbeschreibung ist für die Ursachenanalyse nötig und hilft bei der schnellen Bearbeitung des
Vorgangs. Die Aussage „Defekt“ hilft bei der Fehlersuche
leider nicht.
Please complete this return form to ensure your claim is
processed quickly and efficiently. An accurate description of
the problem is necessary for cause analysis and will help
processing the claim quickly. Unfortunately, stating
“defective” will not help us troubleshoot the issue.
Die RMA-Nummer bekommen Sie von Ihrem
Ansprechpartner im Vertrieb oder Service.
You may obtain the RMA number from your sales or
service representative.
Zu diesem Rücksendeschein gehört eine Dekontaminierungserklärung. Die gesetzlichen Vorschriften schreiben vor, dass
Sie uns diese Dekontaminierungserklärung ausgefüllt und
unterschrieben zurücksenden müssen. Bitte füllen Sie auch
diese im Sinne der Gesundheit unserer Mitarbeiter
vollständig aus.
This return form includes a decontamination statement. The
law requires you to submit this completed and signed
decontamination statement to us. Please complete the entire
form, also in the interest of our employees’ health.
Bringen Sie den Rücksendeschein mit der
Dekontaminierungserklärung bitte zusammen mit den
Versandpapieren in einer Klarsichthülle außen an der
Verpackung an. Ansonsten ist eine Bearbeitung Ihres
Reparaturauftrages nicht möglich!
Attach the return form including decontamination
statement along with the shipping documentation to the
outside of the package, inside a clear pouch. Otherwise
we are unable to process your repair order!
Angaben zum Absender:
Sender information:
Firma / Company
Ansprechpartner /
Contact person
Anschrift / Address
Abteilung /
Department
E-Mai / E-Mail:
Tel. / Phone
Fax / Fax:
Artikelnummer /
Item number
Auftragsnummer /
Order number
Anzahl / Quantity
Rücksendegrund /
Return reason
RMA-Nr. /
RMA no.
Reparatur / Repair
Garantie / Warranty
Zur Prüfung / For inspection
Rückgabe / Return
Vorgangsnummer des Kunden /
Customer transaction number::
Fehlerbeschreibung / Description of the problem:
Ort, Datum /
Place, Date
Seite
Page
1/2
Unterschrift / Stempel /
Signature / Stamp:
Bühler Technologies GmbH
D - 40880 Ratingen, Harkortstr. 29
Tel.: + 49 (0) 2102 / 4989-0 Fax: + 49 (0) 2102 / 4989-20
e-mail: service@buehler-technologies.com
Internet: www.buehler-technologies.com
RMA - Dekontaminierungserklärung
RMA - Decontamination Statement
DE/EN
Gültig ab / valid since: 2014/11/01
Revision / Revision 1
ersetzt Rev. / replaces Rev. 0
Bitte füllen Sie diese Dekontaminierungserklärung für jedes
einzelne Gerät aus.
Gerät / Device
RMA-Nr /
RMA no:
Serien-Nr. / Serial no.
[
]
Please complete this decontamination statement for each
individual item
Ich bestätige hiermit, dass das oben spezifizierte
Gerät ordnungsgemäß gereinigt und dekontaminiert
wurde und keinerlei Gefahren im Umgang mit dem
Produkt bestehen.
I herewith declare that the device as specified above has been
properly cleaned and decontaminated and that there are no
risks present when dealing with the device.
Ansonsten ist die mögliche Gefährdung genauer zu
beschreiben:
In other cases, please describe the hazards in detail:
Aggregatzustand (bitte ankreuzen):
Aggregate state (please check):
Flüssig / Liquid
Fest / Solid
Folgende Warnhinweise sind zu beachten
(bitte ankreuzen):
Pulvrig / Powdery
Please note the following warnings (please check):
Explosiv
Explosive
Giftig / Tödlich
Toxic / lethal
Entzündliche Stoffe
Flammable substances
Komprimierte Gase
Compressed gasses
Gesundheitsgefährdend
Hazardous to health
Gesundheitsschädlich
Harmful to health
Bitte legen Sie ein aktuelles Datenblatt des
Gefahrenstoffes bei!
Ort, Datum /
Place, Date:
Seite
Page
2/2
Gasförmig / Gaseous
Brandfördernd
Oxidizing
Umweltgefährdend
Harmful to the environment
Please include an updated data sheet of the hazardous
substance!
Unterschrift / Stempel
Signature / Stamp:
Bühler Technologies GmbH
D - 40880 Ratingen, Harkortstr. 29
Tel.: + 49 (0) 2102 / 4989-0 Fax: + 49 (0) 2102 / 4989-20
e-mail: service@buehler-technologies.com
Internet: www.buehler-technologies.com