Servomex Functional Safety Manual Servotough OxyExact 2223A Transmitter
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
PROCESS ANALYSERS
SERVOTOUGH OxyExact
2223A Transmitter
Functional Safety Manual
Part Number:
Revision:
Language:
02223006A
0
UK English
This page intentionally blank.
2223A Oxygen Transmitter Functional Safety Manual
LIST OF CONTENTS
Section
1
INTRODUCTION ................................................................................................ 1
1.1
1.2
1.3
Purpose of this manual ....................................................................................... 1
Product description ............................................................................................. 1
Required documentation..................................................................................... 1
2
DEFINITIONS AND DESCRIPTIONS ................................................................ 3
2.1
2.2
2.3
2.4
2.5
2.6
2.7
2.8
2.9
2.10
2.11
2.12
Safety instrumented system (SIS) ...................................................................... 3
Safety function .................................................................................................... 3
Safe failure.......................................................................................................... 3
Dangerous failure................................................................................................ 3
Detected failure................................................................................................... 3
Type A and type B subsystems .......................................................................... 3
Hardware fault tolerance (HFT) .......................................................................... 4
Safe failure fraction (SFF)................................................................................... 4
PFDAVG ................................................................................................................ 4
Mean time to repair (MTTR) ............................................................................... 4
Proof test............................................................................................................. 4
Safety integrity level (SIL) ................................................................................... 5
3
SAFETY INSTRUCTIONS.................................................................................. 7
3.1
3.2
3.3
3.4
3.4.1
3.4.2
3.4.3
3.4.4
3.5
3.6
3.6.1
3.6.2
3.6.3
Boundary of safety system ................................................................................. 7
Safety function .................................................................................................... 7
Installation........................................................................................................... 7
Settings ............................................................................................................... 8
mA Output........................................................................................................... 8
Fault relay ........................................................................................................... 9
Flow alarm .......................................................................................................... 9
Analog inputs ...................................................................................................... 9
Protection of settings ........................................................................................ 10
Checking safety ................................................................................................ 10
Checks where pressure compensation is not used. ......................................... 10
Checks where pressure compensation is used. ............................................... 11
Routine calibration ............................................................................................ 11
A1
APPENDIX........................................................................................................ 13
02223006A / Revision 0
i
2223A Oxygen Transmitter Functional Safety Manual
This page intentionally blank.
ii
02223006A / Revision 0
2223A Oxygen Transmitter Functional Safety Manual
1
INTRODUCTION
1.1
Purpose of this manual
This manual provides information and instructions that will be needed in order to use the
2223A Oxygen Transmitter in a safety instrumented system.
It is aimed at those responsible for planning, designing, installing, commissioning and
maintaining safety instrumented systems using the 2223A Oxygen Transmitter.
1.2
Product description
The 2223A Oxygen Transmitter provides a stable and accurate measurement of the oxygen
concentration in a sample gas stream. Pressure compensation and flow alarm are available
as internally fitted options or as inputs from external measurements.
The measurement is provided as a mA output. Fault indications are provided by a solid
state relay output and by an out-of-range current from the mA output. Additional status
indications are provided by further solid state relays.
The 2223A Oxygen Transmitter is normally used in conjunction with a 2210 or 2213 Control
Unit. This provides additional system inputs and outputs and allows the transmitter to be
configured and measurements to be displayed.
1.3
Required documentation
This manual covers only the functional safety-related aspects of the 2223A Oxygen
Transmitter. For complete information it must be read in conjunction with the following:Title
Part Number
Language
Operator Manual
02210001A
02210011A
02210021A
02223005A
02223015A
02223025A
English
French
German
English
French
German
Installation Manual
02223006A / Revision 0
1
2223A Oxygen Transmitter Functional Safety Manual
This page intentionally blank.
2
02223006A / Revision 0
2223A Oxygen Transmitter Functional Safety Manual
2
DEFINITIONS AND DESCRIPTIONS
2.1
Safety instrumented system (SIS)
A safety instrumented system comprises components and subsystems (from sensor
through to final element) which provide functions aimed at mitigating dangerous failure
conditions.
2.2
Safety function
A defined function executed by a safety instrumented system which is intended to mitigate
a specific dangerous event.
2.3
Safe failure
A failure other than the specific failures which lead to the defined hazardous failure state.
2.4
Dangerous failure
A specific failure mode which leads to a defined non-functioning state. Note that there may
be more than one dangerous failure mode for any particular system.
2.5
Detected failure
A failure that is detected by diagnostic tests, proof tests, operator intervention or through
normal operation.
2.6
Type A and type B subsystems
This relates to the complexity of the subsystem and the degree to which it is understood.
A subsystem is classified as Type A if all of the following are true:•
The failure modes of all components are well defined
•
The behaviour of the subsystem under fault conditions can be completely
determined
•
There is sufficient dependable field failure data to show that the claimed failure
rates for detected and undetected dangerous failures are met.
A subsystem is classified as Type B if any of the following are true:•
The failure mode of at least one component is not well defined
•
The behaviour of the subsystem under fault conditions cannot be completely
determined
•
There is insufficient dependable field failure data to support claims for failure rates
for detected and undetected dangerous failures.
Type B subsystems include those with complex components (e.g. microprocessors) or
software.
02223006A / Revision 0
3
2223A Oxygen Transmitter Functional Safety Manual
2.7
Hardware fault tolerance (HFT)
The ability to continue to provide a safety function in the presence of faults and errors. A
hardware fault tolerance of N means that N+1 faults could cause the loss of a safety
function.
2.8
Safe failure fraction (SFF)
The fraction of failures that does not have the potential to put the safety instrumented
system in a dangerous or non-functioning state. In other words, for each failure mode the
ratio of the diagnosed dangerous failures and the safe failures to the total failures. This
figure is sometimes expressed as a percentage.
2.9
PFDAVG
The probability of a dangerous failure on demand. This is intended to apply to systems
operating in a low demand mode where the safety function is required on average a
maximum of once per year.
A typical safety instrumented system normally consists of three subsystems as follows:-
Sensor
Logic System
Final Elements
e.g. Analyser
e.g. PLC
e.g. Valve
It can be seen that the analyser will typically be a part of the overall system and is therefore
only entitled to a fraction of the total PFDAVG range associated with the specified SIL.
2.10
Mean time to repair (MTTR)
This is the average time taken to restore the safety instrumented system to a working state.
2.11
Proof test
Periodic tests that are performed to detect failures so that, if necessary, the system can be
restored to a fully working state.
4
02223006A / Revision 0
2223A Oxygen Transmitter Functional Safety Manual
2.12
Safety integrity level (SIL)
The international standard IEC 61508 defines four Safety Integrity Levels from SIL1 to
SIL4. Each of these corresponds to a range of probabilities that the safety function will fail.
The higher the SIL the greater the probability that the safety function will work when
required to do so.
The achievable SIL is determined by a number of factors that include the safety
management procedures and lifecycle activities carried out during the development of a
product or system. This manual only considers product hardware failures and so covers the
following characteristics:•
Product type (A or B)
•
Hardware fault tolerance
•
Safe failure fraction
•
Average probability of a dangerous failure of the safety function on demand
(PFDAVG) and associated proof test interval
The following table shows the relationship between hardware fault tolerance and safe
failure fraction for a Type B subsystem (see IEC 61508 Section 2).
SFF
<60%
60 to 90%
90 to 99%
>99%
0
Not allowed
SIL1
SIL2
SIL3
HFT
1
SIL1
SIL2
SIL3
SIL4
2
SIL2
SIL3
SIL4
SIL4
The following table shows the dependency of the SIL on the probability of failure on
demand (PFDAVG) for low demand mode of operation.
Safety Integrity
Level
Low Demand Mode of Operation
Average Probability of Failure to Perform
Safety Function on Demand
4
≥ 10-5 to 10-4
3
≥ 10-4 to 10-3
2
≥ 10-3 to 10-2
1
≥ 10-2 to 10-1
02223006A / Revision 0
5
2223A Oxygen Transmitter Functional Safety Manual
This page intentionally blank.
6
02223006A / Revision 0
2223A Oxygen Transmitter Functional Safety Manual
3
SAFETY INSTRUCTIONS
3.1
Boundary of safety system
The following aspects of the 2223A Oxygen Transmitter are considered to be within the
boundary of the safety system covered in this manual:
•
Paramagnetic cell
•
All signal processing and support functions. This includes optional pressure
compensation and flow alarm when fitted and enabled.
•
mA output
•
“Fault” relay
•
External analog inputs
•
External flow sensor inputs
The following are considered to be outside of the safety boundary and are not covered in
this manual:
3.2
•
Communications interface to Control Unit
•
“Maintenance required” and “Service in Progress” relays
•
Digital inputs
Safety function
The safety function of the 2223A Oxygen Transmitter is the measurement of oxygen
concentration in a sample stream. The dangerous failure is the failure to indicate this
oxygen concentration to the performance stated in the product’s literature.
3.3
Installation
The 2223A Oxygen Transmitter must be installed as detailed in the Installation Manual.
To ensure that the safety instrumented system can respond to detected failures in the
transmitter, the equipment to which the mA output is connected must treat a current of 0mA
as a fault indication.
The fault relay can provide an additional method of notifying fault conditions. However, its
use in safety instrumented systems as the only method of fault indication is not
recommended as it results in a small decrease in the safe failure fraction. If the fault relay is
used then connections should be made to the contacts that are normally open when deenergised.
02223006A / Revision 0
7
2223A Oxygen Transmitter Functional Safety Manual
3.4
Settings
After installation and commissioning according to the installation manual, the following
settings should be made to support the safety function. This will require the use of a 2210
or 2213 Control Unit or a 2215 Tx Interface. To access these functions press the Enter key,
navigate to
Passwords → Enter Password
and enter the appropriate password to allow settings to be changed.
From the initial menu screen, navigate as follows:Transmitters → Transmitter Setup → Select Transmitter → Peripherals
Enter the appropriate forms and ensure that the necessary settings are made as described
below. Note that it may be necessary to scroll the forms displayed on the control unit in
order to see all of the fields shown in the following sections.
3.4.1 mA Output
Configure the mA output as shown as described below:
8
•
Measurement 1 and 2 should be set to appropriate values for the application.
•
Under Range Current must be a non-zero value.
•
Jam must be set to Low or High. If the equipment to which the mA output is
connected does not recognise both 0mA and 21mA as an indication of fault, then
this parameter must be set to Low.
•
On Service In Progress should be set as required.
02223006A / Revision 0
2223A Oxygen Transmitter Functional Safety Manual
3.4.2 Fault relay
Select the Fault Relay and configure as follows.
3.4.3 Flow alarm
If the internal flow alarm is fitted and enabled, it should be calibrated at “normal” gas flow as
described in the operator manual. Alarm levels should then be set as appropriate.
As an alternative to the internal flow sensor, external devices may be used in conjunction
with the external flow alarm inputs as described in the installation manual.
3.4.4 Analog inputs
External measurements may be used to provide pressure compensation and/or cross
interference compensation by connection to the transmitter’s analog inputs. Where this
feature is enabled it should be configured as described below.
02223006A / Revision 0
9
2223A Oxygen Transmitter Functional Safety Manual
3.5
•
Measurement 1, Current 1, Measurement 2 and Current 2 should be configured to
map the input current range to the corresponding measurement range. Note that
Current 1 must be greater than zero.
•
Under Range Current should be set to a value less than Current 1 that represents a
fault condition in the external measurement.
•
Over Range Current should be set to a value greater than Current 2 that represents
a fault condition in the external measurement.
•
Name, Filter Factor and Units should be set as appropriate.
Protection of settings
The 2223A Oxygen Transmitter is supplied with all passwords set to default values. After
configuration these should be altered to provide protection against unauthorised changes.
3.6
Checking safety
Safety checks should be carried out after installation and at regular intervals in accordance
with IEC 61508. The required interval between tests will depend on the application, the
design of the overall safety instrumented system and the target SIL. The recommended
proof test interval is a maximum of one year.
If there are any transmitter fault conditions present, these must be resolved before
proceeding. It is also recommended that any maintenance-required conditions are
addressed.
For all tests it is assumed that the 2223A Oxygen Transmitter is powered and has reached
stable operating temperature.
3.6.1 Checks where pressure compensation is not used.
10
1.
Take appropriate action to ensure that spurious outputs will not compromise safety or
cause false trips.
2.
Using the built-in analog output test feature, set the output to 0.0mA and verify that
the correct current is generated. Repeat with the output set to 20.0mA.
3.
Carry out a full low and high calibration.
4.
Perform a reference measurement with an oxygen concentration between the low
and high calibration points. Check the mA output current and verify that it is within
the required tolerance.
5.
If an internal flow sensor is fitted, alter the flow of gas through the analyser and
check that the flow alarms are triggered at the appropriate levels.
6.
Restore normal operation.
02223006A / Revision 0
2223A Oxygen Transmitter Functional Safety Manual
3.6.2 Checks where pressure compensation is used.
1.
Take appropriate action to ensure that spurious outputs will not compromise safety or
cause false trips.
2.
Using the built-in analog output test feature, set the output to 0.0mA and verify that
the correct current is generated. Repeat with the output set to 20.0mA.
3.
Navigate to the Transmitter Diagnostics form and view Calibrated Pressure. Check
that the reading displayed is within an acceptable tolerance of the actual pressure.
4.
Calibrate the pressure sensor.
5.
Carry out a full low and high calibration of the oxygen measurement.
6.
Carry out pressure compensation calibration.
7.
Perform a reference measurement with an oxygen concentration between the low
and high calibration points. Check the mA output current and verify that it is within
the required tolerance.
8.
If an internal flow sensor is fitted, alter the flow of gas through the analyser and
check that the flow alarms are triggered at the appropriate levels.
9.
Restore normal operation.
3.6.3 Routine calibration
Regular full calibrations should be carried out as recommended in the operator manual.
02223006A / Revision 0
11
2223A Oxygen Transmitter Functional Safety Manual
This page intentionally blank.
12
02223006A / Revision 0
2223A Oxygen Transmitter Functional Safety Manual
A1
APPENDIX
02223006A / Revision 0
13