User’s
Manual
Model GD402
Gas Density Meter
IM 11T03E01-01E
IM 11T03E01-01E
11th Edition (YK)
i
Introduction
The manual is applied to the following product.
Detector:
GD40G (General Purpose), GD40R (TIIS) (Style S2),
GD40T (FM), GD40V (CSA) (Style S1)
Converter:
GD402G (General Purpose), GD402R (TIIS),
GD402T (FM), GD402V (CSA) (Style S2)
Note:
When using GD40/402 as CE marking compliance product, select GD40G and GD402G.
The related documents are as follows.
General Specifications: GS 11T3E1-01E
User’s Manual:
IM 11T03E01-01E (this manual)
* the “E” in the document number is the language code.
An exclusive User’s Manual might be attached to the products whose suffix codes or option
codes contain the code “Z” (made to customers’ specifications). Please read it along with this
manual.
The GD402 gas density meter and the GD40 detector not only provide continuous measurement
of gas density, but also several other valuable parameters, including specific gravity and
molecular weight. The GD40 detector is designed for intrinsically safe and flame-proof,
explosion protected applications. It is designed to be virtually maintenance free for all accepted
applications.
The GD402 is a rugged microprocessor-based converter designed in two versions to meet both
general area and explosion-proof application requirements. In addition to the display of several
key data items, the converter also provides the choice of three different means for calibration:
automatic; semi-automatic and one-touch manual operation.
l Cross-Checking the Specifications
CAUTION
The GD402 gas density meter is shipped after adjusting the both of detector and converter in
pairs. When installation, confirm whether the serial number on both of converter and detector are
in pairs or not. If mismatched in pairs, converter is to be out of order. When converter or detector
supplied individually, enter the detector constants, described on inside the lid of the GD40, into
converter so that the GD402 is going to be well.
In detail, please refer to Figure 5.30, Figure 6.31 or Figure 7.28.
Upon delivery of the purchased product, unpack it carefully and make sure it is completely free
from damage that may have occurred during transport. It must be shipped in strict conformance
to the purchaser’s specifications. By way of precaution, confirm that the equipment is the exact
model you ordered. Also check that all accessory components are included, when confirming the
specifications, refer to the model and suffix codes indicated on the nameplate on the equipment.
For a description of the model and suffix codes, refer to Chapter 1,’’Specifications.’’
l Consideration of Operation Parameters
The GD402 operates with the same parameters set when it was delivered (default data) , when it
is put into operation under these conditions.
Media No. IM 11T03E01-01E
11th Edition : June. 2017 (YK)
All Rights Reserved Copyright © 1998, Yokogawa Electric Corporation
IM 11T03E01-01E
ii
Before starting measurement, check whether or not the default data meets your operating
conditions. If necessary, re-set the parameters to suit your operating requirements.
To check the defaults, make use of the sheet, “Records of GD402’s Operation Parameter
Settings,” in the back of this manual. It is advisable that, if any of the operation parameter settings
have been changed, the new data be noted in this record.
l Information Covered in This Manual
This manual covers all of the information for handling the GD402 converter and the GD40
detector, including instructions on installation, operation, setting of operation parameters,
inspection and maintenance.
l Equipments
The GD402 should only be used with equipment that meets the relevant IEC, American,
Canadian standard. Yokogawa accepts no responsibility for the misuse of this unit.
CAUTION
Modification and adaptation of the instrument’s parts, wiring and construction are prohibited.
WARNING
Power Supply
Ensure the power supply voltage matches instrument before turning ON the power.
Protective grounding
Make sure to connect the protective grounding to prevent an electric shock before turning
ON the power.
l Necessity of protective grounding
Never cut off the internal or external protective grounding wire or disconnect the wiring of
protective grounding terminal. Doing so poses a potential shock hazard.
l Defect of Protective Grounding and Fuse
Do not operate the instrument when protective grounding or fuse might be defective.
l Fuse
To prevent a fire, make sure to use fuses with specified standard (voltage, current, type).
Before replacing the fuses, turn off the power and disconnect the power source.
Do not use a different fuse or short-circuit the fuse holder.
l Do not Remove any Covers
There are some areas with high voltages. Do not remove any cover if the power supply is
connected. The cover should be removed by qualified personnel only.
l
l
IM 11T03E01-01E
iii
n Safety, Protection, and Modification of the Product
• In order to protect the system controlled by the product and the product itself and ensure
safe operation, observe the safety precautions described in this user’s manual. We assume
no liability for safety if users fail to observe these instructions when operating the product.
• If this instrument is used in a manner not specified in this user’s manual, the protection
provided by this instrument may be impaired.
• If any protection or safety circuit is required for the system controlled by the product or for
the product itself, prepare it separately.
• Be sure to use the spare parts approved by Yokogawa Electric Corporation (hereafter
simply referred to as YOKOGAWA) when replacing parts or consumables.
• Modification of the product is strictly prohibited.
• The following safety symbols are used on the product as well as in this manual.
WARNING
This symbol indicates that an operator must follow the instructions laid out in this manual in order
to avoid the risks, for the human body, of injury, electric shock, or fatalities. The manual describes
what special care the operator must take to avoid such risks.
CAUTION
This symbol indicates that the operator must refer to the instructions in this manual in order to
prevent the instrument (hardware) or software from being damaged, or a system failure from
occurring.
CAUTION
This symbol gives information essential for understanding the operations and functions.
NOTE
This symbol indicates information that complements the present topic.
This symbol indicates Protective Ground Terminal.
This symbol indicates Function Ground Terminal. Do not use this terminal as the
protective ground terminal.
Alternating current
Direct current
[Notational Conventions Specific to This Manual]
The following notational conventions apply to the representations of operation keys, information
shown on the display, and information indicated on the product itself when they are discussed
specifically in the text of this manual.
l Operation Keys
Indicated with brackets [ ] or " " as: [YES] Key or "YES" key
l Information Shown on Display
Indicated with braces [ ] as:
[HOLD], meaning the status
[YES], meaning the indicator on an operation key
[CALIB], meaning a message
[205] (lit) or [205] (blinking), meaning the data item shown along with its state
IM 11T03E01-01E
iv
l Information Indicated on Product
Indicated with angle brackets < > as: (lit) or (unlit), meaning the contact output indicator
lamp along with its state
mode, meaning the mode of measurement
l Information on the State of Blinking
Indicated in shaded typography as:
(blinking)
, in contrast to (lit)
IM 11T03E01-01E
v
CE marking products
u
n Identification Tag
This manual and the identification tag attached on packing box are essential parts of the product.
Keep them together in a safe place for future reference.
n Users
This product is designed to be used by a person with specialized knowledge.
n How to dispose the batteries
This is an explanation about the new EU Battery Directive (DIRECTIVE 2006/66/EC). This
directive is only valid in the EU.
Batteries are included in this product. Batteries incorporated into this product cannot be removed
by yourself. Dispose them together with this product.
When you dispose this product in the EU, contact your local Yokogawa Europe B.V. office. Do not
dispose them as domestic household waste.
Battery type: Manganese dioxide lithium battery
Notice : The symbol (see above) means they shall be sorted out and collected as ordained in ANNEX II in DIRECTIVE 2006/66/EC.
n Product Disposal
The instrument should be disposed of in accordance with local and national legislation/
regulations.
n Authorized Representative in EEA
The Authorized Representative for this product in EEA is Yokogawa Europe B.V. (Euroweg 2,
3825 HD Amersfoort, The Netherlands).
IM 11T03E01-01E
u
vi
After-sales Warranty
n Do not modify the product.
n Yokogawa warrants the product for the period stated in the pre-purchase
quotation Yokogawa shall conduct defined warranty service based on its
standard. When the customer site is located outside of the service area, a fee
for dispatching the maintenance engineer will be charged to the customer.
n During the warranty period, for repair under warranty carry or send the
product to the local sales representative or service office. Yokogawa will
replace or repair any damaged parts and return the product to you.
l Before returning a product for repair under warranty, provide us with the model name
and serial number and a description of the problem. Any diagrams or data explaining the
problem would also be appreciated.
l If we replace the product with a new one, we won’t provide you with a repair report.
n In the following cases, customer will be charged repair fee regardless of
warranty period.
• Failure of components which are out of scope of warranty stated in instruction manual.
• Failure caused by usage of software, hardware or auxiliary equipment, which Yokogawa
Electric did not supply.
• Failure due to improper or insufficient maintenance by user.
• Failure due to modification, misuse or outside-of-specifications operation which Yokogawa
does not authorize.
• Failure due to power supply (voltage, frequency) being outside specifications or abnormal.
• Failure caused by any usage out of scope of recommended usage.
• Any damage from fire, earthquake, storms and floods, lightning, disturbances, riots, warfare,
radiation and other natural changes.
n Yokogawa does not warrant conformance with the specific application at the
user site. Yokogawa will not bear direct/indirect responsibility for damage due
to a specific application.
n Yokogawa Electric will not bear responsibility when the user configures the
product into systems or resells the product.
n Maintenance service and supplying repair parts will be covered for five years
after the production ends. For repair for this product, please contact the
nearest sales office described in this instruction manual.
IM 11T03E01-01E
Toc-1
Model GD402
Gas Density Meter
IM 11T03E01-01E 11th Edition
CONTENTS
Introduction...............................................................................................................i
u
CE marking products ...................................................................................v
u
After-sales Warranty....................................................................................vi
1. SPECIFICATIONS...................................................................................... 1-1
1.1
1.2
1.3
2.
Specifications..................................................................................................... 1-1
1.1.1
General Specifications........................................................................ 1-1
1.1.2
GD40G, T, V, R Detector..................................................................... 1-4
1.1.3
GD402G, T, V, R Converter................................................................ 1-4
Model and Suffix Codes.................................................................................... 1-6
1.2.1
Gas Density Converter....................................................................... 1-6
1.2.2
Gas Density Detector.......................................................................... 1-6
1.2.3
Bracket for Connection with External Cables
(For Explosion-proof use)................................................................... 1-7
1.2.4
Two-core, Double-Shielded Cable...................................................... 1-7
1.2.5
Brain Terminal (Optional).................................................................... 1-8
1.2.6
Pressure transmitter (Optional).......................................................... 1-8
External Views and Dimension......................................................................... 1-9
1.3.1
GD402G Converter (General purpose).............................................. 1-9
1.3.2
Pipe- and Wall-Mounting Bracket (Optional).................................... 1-10
1.3.3
GD402T, V, R Converter (Explosion-proof)...................................... 1-11
1.3.4
GD40 Detector.................................................................................. 1-12
1.3.5
Detector Unit (Intrinsically Safe, Explosion-proof)............................ 1-13
INSTALLATION, WIRING AND PIPING..................................................... 2-1
2.1
2.2
2.3
Installing the Detector....................................................................................... 2-1
2.1.1
Selecting the Location......................................................................... 2-1
2.1.2
GD40T (FM Explosion-proof and Intrinsically Safe Approval)............ 2-2
2.1.3
GD40V (CSA Explosion-proof and Intrinsically Safe Approval)......... 2-3
2.1.4
Mounting the Detector........................................................................ 2-4
Installing the Converter..................................................................................... 2-5
2.2.1
Selecting the Location......................................................................... 2-5
2.2.2
GD402T (FM Explosion-proof Approval)............................................ 2-6
2.2.3
GD402V (CSA Explosion-proof Approval).......................................... 2-7
2.2.4
Mounting the Converter...................................................................... 2-8
Piping................................................................................................................ 2-10
IM 11T03E01-01E
Toc-2
2.4 Wiring................................................................................................................ 2-12
2.4.1
Wiring Procedure.............................................................................. 2-12
2.4.2
Cables Wired to Power Supply......................................................... 2-16
2.4.3
Cables Wired to Outputs................................................................... 2-17
2.4.4
Cables Wired to Contact I/Os........................................................... 2-18
2.4.5
Cable Wired to GD40 Detector...................................................... 2-19
2.4.6
Wiring to Model EJX310A pressure transmitter................................ 2-20
2.4.7
Cables Wired to the Ground............................................................. 2-20
3. OPERATION............................................................................................... 3-1
3.1
3.2
3.3
Preparation for Operation................................................................................. 3-1
3.1.1
Inspecting Installation, Piping and Wiring Workmanship................... 3-1
3.1.2
Supplying Power................................................................................. 3-2
3.1.3
Display on Operation Panel and Operation Keys............................... 3-3
3.1.4
Basic Key Operation........................................................................... 3-4
3.1.5
Checking the Setting Parameters ...................................................... 3-6
3.1.6
Calibrating the Analyzer for Correct Readings................................... 3-6
3.1.7
Checking the Analyzer for Performance............................................. 3-7
Normal Operation............................................................................................... 3-8
3.2.1
Starting Operation............................................................................... 3-8
3.2.2
Corrective Actions in Case of Failure.................................................. 3-8
3.2.3
Inspection and Maintenance............................................................... 3-9
Shutdown and Restart....................................................................................... 3-9
3.3.1
Measures for Shutdown...................................................................... 3-9
3.3.2
Measures for Restarting..................................................................... 3-9
4. FUNCTIONS............................................................................................... 4-1
4.1
4.2
5.
4.1.1
Measurement, Operation, Setting and Service.................................. 4-1
4.1.2
Key Operations................................................................................... 4-1
4.1.3
Points to Be Noted When Making Settings......................................... 4-2
Setting Lists........................................................................................................ 4-2
DENSITY / PARAMETER SETTING.......................................................... 5-1
5.1
5.2
6.
Summary of Setting Operations....................................................................... 4-1
Setting Parameters............................................................................................ 5-2
5.1.1
Setting Parameters at Measurement Level........................................ 5-2
5.1.2
Setting Parameters at Operation Level.............................................. 5-3
5.1.3
Setting Parameters at Setting Level................................................... 5-4
5.1.4
Setting Parameters at Service Level.................................................. 5-5
Parameter Setting.............................................................................................. 5-6
5.2.1
Parameter Setting at Operation Level................................................ 5-6
5.2.2
Parameter Setting at Setting Level................................................... 5-10
5.2.3
Parameter Setting at Service Level.................................................. 5-16
CALORIE / PARAMETER SETTING......................................................... 6-1
6.1
Setting Parameters............................................................................................ 6-2
IM 11T03E01-01E
6.2
7.
Toc-3
6.1.1
Setting Parameters at Measurement Level........................................ 6-2
6.1.2
Setting Parameters at Operation Level.............................................. 6-3
6.1.3
Setting Parameters at Setting Level................................................... 6-4
6.1.4
Setting Parameters at Service Level.................................................. 6-5
Parameter Setting.............................................................................................. 6-7
6.2.1
Parameter Setting at Operation Level................................................ 6-7
6.2.2
Parameter Setting at Setting Level................................................... 6-12
6.2.3
Parameter Setting at Service Level.................................................. 6-16
HYDROGEN PURITY/PARAMETER SETTING........................................ 7-1
7.1
7.2
8.
Setting Parameters............................................................................................ 7-2
7.1.1
Setting Parameters at Measurement Level........................................ 7-2
7.1.2
Setting Parameters at Operation Level.............................................. 7-2
7.1.3
Setting Parameters at Setting Level................................................... 7-3
7.1.4
Setting Parameters at Service Level.................................................. 7-4
Parameter Setting.............................................................................................. 7-6
7.2.1
Parameter Setting at Measurement Level.......................................... 7-6
7.2.2
Parameter Setting at Operation Level................................................ 7-7
7.2.3
Parameter Setting at Setting Level................................................... 7-10
7.2.4
Parameter Setting at Service Level.................................................. 7-14
DENSITY /CALIBRATION PROCEDURE................................................. 8-1
8.1
Basic Calibration Procedure............................................................................. 8-1
8.2
Preparation for Calibration............................................................................... 8-1
8.2.1
Setting Type of Calibration and Checking Valve Operation................ 8-1
8.2.2
Setting Calibration Data...................................................................... 8-3
8.3 Calibration.......................................................................................................... 8-4
9.
8.3.1
Semi-automatic Calibration................................................................ 8-4
8.3.2
Manual Calibration.............................................................................. 8-5
8.3.3
Automatic Calibration.......................................................................... 8-6
CALORIE /CALIBRATION PROCEDURE................................................. 9-1
9.1
Basic Calibration Procedure............................................................................. 9-1
9.2
Preparation for Calibration............................................................................... 9-1
9.2.1
Setting Type of Calibration and Checking Valve Operation................ 9-1
9.2.2
Setting Calibration Data...................................................................... 9-3
9.3 Calibration.......................................................................................................... 9-5
10.
9.3.1
Semi-automatic Calibration................................................................ 9-5
9.3.2
Manual Calibration.............................................................................. 9-6
9.3.3
Automatic Calibration.......................................................................... 9-7
HYDROGEN PURITY /CALIBRATION PROCEDURE........................... 10-1
10.1
Basic Calibration Procedure........................................................................... 10-1
10.2
Preparation for Calibration............................................................................. 10-1
10.3 Calibration........................................................................................................ 10-2
IM 11T03E01-01E
11.
Toc-4
INSPECTION AND MAINTENANCE....................................................... 11-1
11.1
Routine Inspection and Maintenance............................................................ 11-1
11.1.1
Checking Readings and Calibrating the Analyzer............................ 11-1
11.1.2
Checking the Flowrate of Sample Gas............................................. 11-1
11.1.3
Periodic Replacement of the Detector's O-rings.............................. 11-1
11.1.4
Replacing the Fuse........................................................................... 11-2
11.1.5 Cleaning............................................................................................ 11-2
11.2
Inspection In Case of Failure.......................................................................... 11-3
11.2.1
Inspecting the Analyzer in an Alarm Status...................................... 11-3
11.2.2
Inspecting the Analyzer in a FAIL Status.......................................... 11-4
Customer Maintenance Parts List............................................CMPL 11T3E1-01E
Customer Maintenance Parts List............................................CMPL 11T3E1-02E
Customer Maintenance Parts List............................................CMPL 11T3E1-03E
Revision Information................................................................................................i
IM 11T03E01-01E
1-1
<1. Specifications>
1. SPECIFICATIONS
1.1
Specifications
The specifications of the GD402 gas density meter as described below.
Certificate is approved for model GD402G, GD40G.
1.1.1
General Specifications
l System Components
(1) GD40G, T, V, R Detector: Rainproof for outdoor use (equivalent to IP65/NEMA4X)
(see note under “1.1.2 Ambient conditions” on page 1-5.)
Ambient Temperature : -10 to 60°C
Ambient Humidity :
5 to 95%RH
GD40G :
General purpose detector. (Non-Explosion-proof)
Electrical connection : 1/2 NPT female
Process connection : 1/4 NPT female
GD40T :
FM Explosion-proof and Intrinsically safe Approval.
Explosion-proof for Class I, Division 1, Groups B, C and D;
Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G with
Intrinsically Safe sensor for Class I, II, III, Division 1, Groups B, C, D, E, F
and G.
Applicable standards:
FM Class 3600:1998, FM Class 3610:2010, FM Class 3615:1989, FM Class
3810:2005, ANSI/ISA 60079-0:2009, ANSI/UL 60079-11:2009, ANSI/NEMA 250:1991
Enclosure :
NEMA Type 4X
Temperature Code :
T5
Electrical connection : 1/2 NPT female
Process connection : 1/4 NPT female
GD40V :
CSA Explosion-proof and Intrinsically safe Approval.
Explosion-proof for Class I, Division 1, Groups B, C and D;
Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G with Intrinsically
Safe sensor for Class I, II, III, Division 1, Groups B, C, D, E, F and G.
Applicable standards:
CAN/CSA-C22.2 No. 25-1966, CAN/CSA-C22.2 No. 30-M1986, CAN/CSA-C22.2
No. 94.2-15, CAN/CSA-C22.2 No. 157-92, CAN/CSA-C22.2 No. 1010.1-92
Enclosure :
Type 4X
Temperature Code :
T5
Electrical connection : 1/2 NPT female
Process connection : 1/4 NPT female
GD40R :
TIIS Explosion-proof and Intrinsically safe Approval.
Explosion-proof code : Exd [ia] IIB+H2T5
Temperature Code :
T5
Electrical connection : G1/2 female
Process connection : Rc1/4
IM 11T03E01-01E
<1. Specifications>
1-2
(2) GD402G, T, V, R Converter : Rainproof for outdoor use (equivalent to IP65 / NEMA 4X)
Ambient Temperature :
-10 to 55°C
Ambient Humidity :
5 to 95%RH
GD402G :
General purpose converter. (Non-Explosion-proof).
Electrical connection: 21mm (0.9inch) in diameter. Pg 13.5 cable glands included
GD402T : FM Explosion-proof Approval.
Explosion-proof for Class I, Division 1, Groups B, C and D;
Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G.
Applicable standards:
FM Class 3600:1998, FM Class 3615:1989, FM Class 3810:1989, NEMA 250:1991
Enclosure :
NEMA Type 4X
Temperature Code :
T6
Electrical connection : 1/2 NPT female
GD402V : CSA Explosion-proof Approval.
Explosion-proof for Class I, Division 1, Groups B, C and D;
Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G.
Applicable standards:
CAN/CSA-C22.2 No. 0-M91, CAN/CSA-C22.2 No. 0.4-M1982, CAN/CSA-C22.2
No. 0.5-1982, CAN/CSA-C22.2 No. 25-1966, CAN/CSA-C22.2 No. 30-M1986,
CAN/CSA-C22.2 No. 94-M91, CAN/CSA-C22.2 No. 1010.1-92
Enclosure :
Type 4X
Temperature Code : T6
Electrical connection : 1/2 NPT female
GD402R :
TIIS Explosion-proof Approval.
Explosion-proof code : Exd IIB+H2T6
Temperature Code : T6
Electrical connection : G3/4 female
(3) EJX310A Absolute pressure transmitter (optional)
See GS 01C25D01-01EN for EJX310A.
l Sample Gas conditions
Sample gas: All gases except for corrosive gas and acetylene gas.
Temperature: -10 to 50°C (non-condensing)
Pressure:
Max. 588.5 kPa (abs)
Gas flow:
0.1 to 1 L/min
l Output Signals
Output 1:
4-20 mA DC
Isolated from inputs; load resistance: 600 Ω maximum (Load resistance of 250550 Ω required when in the BRAIN communication mode.)
Output 2:
4-20 mA DC
Isolated from inputs; load resistance: 600 Ω maximum
l Power Supply
Rated voltage range:
100 to 240 V AC, 24 V DC
Allowable voltage range:
85 to 264 V AC, 21.6 to 26.4 V DC
Rated frequency:
50 or 60 Hz
Allowable frequency range: 47 to 63 Hz
IM 11T03E01-01E
1-3
<1. Specifications>
l Power Consumption
Approximately 12 W.
l Fuse
250 V 1 A time lag type authorized VDE/SEMKO 100 to 240 V AC Model
250 V 2 A time lag type authorized VDE/SEMKO 24V DC Model
l Safety, EMC, and RoHS conformity standards
Installation Altitude: 2000 m or less
Category based on IEC 61010: II (Note)
Pollution degree based on IEC 61010: 2 (Note)
Note: Installation category, called over-voltage category, specifies impulse withstand voltage.
Category II is for electrical equipment.
Pollution degree indicates the degree of existence of solid, liquid, gas or other inclusions which may reduce dielectric
strength. Degree 2 is the normal indoor environment.
Safety Standards:
EN 61010-1, EN 61010-2-030
*: Applied only when GD402G converter is used with GD40G detector.
EMC Standards:
EN 61326-1* Class A Table 2,
EN 61326-2-3, EN 61000-3-2, EN 61000-3-3
EMC Regulatory Arrangement in Australia and New Zealand (RCM) EN61326-1 Class A
Korea Electromagnetic Conformity Standard**
* : Influence of immunity environment (Criteria A): ±50% of F.S.
**: Applied only when GD402G converter is used with GD40G detector.
**: Applied only when GD402R converter is used with GD40R detector.
RoHS:
EN 50581
l Characteristics
GD402 specification list
Density kg/m3
Density lb/ft3
Specific Gravity Molecular Weight Concentration vol%
0 - 6 (compensated) 0 - 0.4 (compensated)
Range
0-5
0 - 140
0 - 100
0 - 60 (physical) 0 - 4 (physical)
Concentration equivalent
Minimum Range 0.1
0.01
0.1
4
to 100 kg/m3
Response Time 90% approx. 5 sec
approx. 5 sec
approx. 5 sec
approx. 5 sec
approx. 5 sec
Linearity
+/-1% FS
+/-1% FS
+/-1% FS
+/-1% FS
+/- 1
+/- 0.001
+/- 0.0001
+/- 0.001
+/- 0.02
+/- 0.5% or Concentration
Repeatability
or +/-0.5%FS * or +/-0.5%FS*
or +/-0.5%FS*
or +/-0.5%FS*
equivalent to +/-0.001 kg/m3 *
Long term
Concentration equivalent
+/- 0.003/month +/- 0.002/month
+/- 0.003/month +/- 0.07/month
stability
to +/- 0.003 kg/m3/month
Item
*:
Whichever is greater
Density is the basic measurement, the other representations are derived from the Density data.
Item
H2 in Air vol% H2 in CO2 vol% Air in CO2 vol%
Range
85 - 100
0 - 100
0 - 100
Minimum
–
–
–
Range
Response
approx. 5 sec approx. 5 sec approx. 5 sec
Time 90%
Linearity
+/- 1
+/- 1
+/- 1
Repeatability
+/- 0.5
+/- 0.5
+/- 0.5
Drift
+/- 0.5/month +/- 0.5/month +/- 0.5/month
Caloric value MJ/m3
0 - 130
Caloric value equivalent
to 0.100 kg/m3
approx. 5 sec
British Thermal Unit KBTU/ft3
0 - 3.5
Caloric value equivalent
to 0.100 kg/m3
approx. 5 sec
+/- 1%FS
+/- 0.5%FS or Caloric value
equivalent to 0.001kg/m3*
Caloric value equivalent
to +/- 0.003 kg/m3/month
+/- 1%FS
+/- 0.5%FS or Caloric value
equivalent to 0.001 kg/m3 *
Caloric value equivalent
to +/- 0.0025/month
Caloric Value and BTU are possible representations of the Density.
GD402 does not contain table information, only a single mathematical equation.
*: Whichever is greater
IM 11T03E01-01E
<1. Specifications>
1.1.2
1-4
GD40G, T, V, R Detector
l Material exposed to gas
SUS 316 stainless steel, Acrylonitrile Butadiene Rubber and Fluorine-contained Rubber
(O-ring)
l Ambient conditions
Temperature:
-10 to 60°C (14 to 140 °F)
Humidity:
5 to 95%RH
Installation:
Pipe-mounted or on panel
Construction:
Explosion-proof and Intrinsically safe. (See 1.1.1 “System Components”)
hough the detector construction makes it relatively insensitive to sudden changes in
T
the gas temperature, extra precision can be achieved by keeping ambient temperature
conditions as constant as possible. In measurements where optimum precision
is required it is therefore not recommended to install the detector in an outdoors
environment, especially not if such installation is prone to direct sunlight.
l Finish
Cover:
Equivalent to Munsell 0.6GY3.1/2.0
Case:
Equivalent to Munsell 2.5Y8.4/1.2
l Weight
Approximately 7 kg (with pipe mounting bracket)
l Detector unit
When the system is ordered to be used as a hydrogen purity meter, an optional pressure
transmitter is required for pressure compensation.
• If /EJAJ1, /EJAF2, /EJAF3 or, /EJAF4 are ordered, the detector unit and the
pressure transmitter and the tubing in between will all be integrated on a single mounting
plate. This allows the space where the pressure transmitter is normally mounted to be
used effectively for other purposes.
1.1.3
GD402G, T, V, R Converter
l Display
Reading:
Digital (6 digits maximum; resolution: 0.0001 kg/m3)
Data items shown:
Measured value:
Always on display.
Alarm indications: Abnormal concentration, abnormal pressure range of input and abnormal
value of calibration.
Parameters for calibration: Time of calibration, settling time, starting time of calibration and
calibration cycle
Self-diagnostic indications: Sensor oscillation shutdown, abnormal oscillation frequency of
sensor, failure in sensor temperature detection, failure in A/D
conversion stage and memory failure.
Alarm settings: The contact state can be set to either “normally open (NO)” or “normally
closed (NC)” depending on the needs of the application.
Temperature:
Temperature of gas being measured
IM 11T03E01-01E
1-5
<1. Specifications>
l Contact Outputs/Input
Contact output:
Contact input:
Signals for Maintenance, Fail, Hi/Lo alarms
Contact capacity: 250 V AC at 3 A or 30 V DC at 3 A
Signal for switching between the Hydrogen Purity meter and the
Replacement meter
l Calibration
Manual (one touch), Semi automatic, Automatic calibration
l Communication
Protocol:
BRAIN communication
Data items that can be transmitted by the hand-held terminal are
numerical data, such as concentration, temperature and pressure, alarm
setpoint and self-diagnostic parameters.
l Ambient Conditions
Temperature:
Humidity:
-10 to 55°C (14 to 131°F)
5 to 95%RH
l Installation
Non-Explosion-proof models: Pipe-, panel- or wall-mounted
Explosion-proof models:
Pipe- or panel-mounted
l Finish
GD402G (general purpose):
Front cover:
Equivalent to Munsell 0.6GY3.1/2.0
Case:
Equivalent to Munsell 2.5Y8.4/1.2
GD402T, V, R (explosion-proof):
Equivalent to Munsell 0.6GY3.1/2.0
l Weight
GD402G (general purpose): Approx. 3 kg (including mounting bracket)
GD402T, V, R (explosion-proof): Approx. 15 kg (including mounting bracket)
l Fuse
250 V 1 A time lag type authorized VDE/SEMKO 100 to 240 V AC Model
250 V 2 A time lag type authorized VDE/SEMKO 24 V DC Model
IM 11T03E01-01E
1.2
<1. Specifications>
1-6
Model and Suffix Codes
1.2.1
Gas Density Converter
[Style: S2]
Model
GD402G
GD402T
Suffix code Option code
Description
- - - - - - - - - - - - - - - - - - General purpose model, 6 cable glands included.
- - - - - - - - - - - - - - - - - - FM certified explosion-proof model. Gland threads 1/2 NPT. No cable
glands included.
GD402V
- - - - - - - - - - - - - - - - - - CSA certified explosion-proof model. Gland threads 1/2 NPT. No cable
glands included.
GD402R
- - - - - - - - - - - - - - - - - - TIIS certified explosion-proof model. Gland threads G3/4. No cable
glands included.
Power
-D
- - - - - - - - - 24 V DC
supply
-A
- - - - - - - - - 100-240 V AC
Label and
-E
- - - - - - - - - English label
approval
-J
- - - - - - - - - TIIS approval, English label (only GD402R)
Instruction Manual
-E - - - - - - - - - English
Options
/PA
Panel mounting
(only GD402G)
/U
Universal (Pipe and Wall) Mounting
/H
Hood
Note: Explosion-proof models, GD402T, V, R have only pipe mounting bracket as standard.
n GD402
Standard Accessory List
Model
GD402G
Item
Qty
Part Number
Fuse
1
A1512EF (Power Supply: 100-240 V AC)
A1111EF (Power Supply: 24 V DC)
Universal Mount Set
1
K9171SS
Panel Mount Set
1
K9171ST
Fuse
1
A1512EF (Power Supply: 100-240 V AC)
A1111EF (Power Supply: 24 V DC)
1
K9214HD
1
K9214HE
Bracket
GD402R
GD402T Bracket
GD402V U-Bolt Assy
1.2.2
2
D0117XL-A
Bolt
1
Y9835NU
Bolt
2
Y9820NU
Gas Density Detector
[Style: S2]
Model
GD40G
GD40R
Label
approval
Options
Suffix code Option code
Description
- - - - - - - - - - - - - - - - - - General purpose detector. 1/4 NPT gas threads and 1/2 NPT gland
threads. No cable gland included. Mounting bracket included.
- - - - - - - - - - - - - - - - - - TIIS certified explosion - proof detector. Rc1/4 gas threads and G1/2 gland
threads. Cable gland included. Mounting bracket included.
-E
- - - - - - - - - English label, no approval (only GD40G)
-J
- - - - - - - - - TIIS approval, English label (only GD40R)
/EJAJ1
TIIS certified EJX mounted with detector on mounting plate. Rc1/4 gas
threads and G1/2 gland thread. Cable gland included.(only GD40R)
/EJAJ1T
TIIS certified EJX mounted with detector on mounting plate. Rc1/4 gas threads
and G1/2 gland thread. Cable gland included. EJX with TAG (only GD40R)
/EJAF2
EJX mounted with detector on mounting plate.1/4 NPT gas threads and
1/2 NPT gland threads. No cable gland included.(only GD40G)
/EJAF2T
EJX mounted with detector on mounting plate.1/4 NPT gas threads and 1/2
NPT gland threads. No cable gland included. EJX with TAG (only GD40G)
IM 11T03E01-01E
1-7
<1. Specifications>
[Style: S1]
Model
GD40T
GD40V
Options
Suffix code Option code
Description
- - - - - - - - - - - - - - - - - - FM certified explosion - proof detector. 1/4 NPT gas threads and 1/2 NPT
gland thread. No cable gland included. Mounting bracket included
- - - - - - - - - - - - - - - - - - CSA certified explosion - proof detector. 1/4 NPT gas threads and 1/2 NPT
gland thread. No cable gland included. Mounting bracket included.
/EJAF3
FM certified EJX mounted with detector on mounting plate.1/4 NPT gas
threads and 1/2 NPT gland thread. No cable gland included. (only GD40T)
/EJAF3T
FM certified EJX mounted with detector on mounting plate.1/4 NPT gas threads
and 1/2 NPT gland thread. No cable gland included. EJX with TAG (only GD40T)
/EJAF4
CSA certified EJX mounted with detector on mounting plate. 1/4 NPT gas
threads and 1/2 NPT gland thread. No cable gland included. (only GD40V)
/EJAF4T
CSA certified EJX mounted with detector on mounting plate. 1/4 NPT gas threads
and 1/2NPT gland thread. No cable gland included. EJX with TAG (only GD40V)
n GD40 Standard Accessory
Item
Qty
Part Number
U-Bolt Assy *1
4
D0117XL-A
Bracket *1
1
K9214HD
Bracket *1
1
K9214HE
Gland *2
1
G9601AM
*1:
*2:
Not supplied when option code “/EJAJ1”, “/EJAF2”, “/EJAF3” or “/EJAF4” is specified.
Supplied only for GD40R.
1.2.3
Bracket for Connection with External Cables
(For Explosion-proof use)
Part No.
L9811LL
Description
G3/4 explosion-proof cable gland.
Cable’s outside diameter 8 to 16 mm.
Used for the GD402R converter.
Note: Specify the number of cable glands for converter in hazardous area
1.2.4
Two-core, Double-Shielded Cable
Normally two conductor shielded cable can be used, but when failure arises from noises
disturbance, this cable is recommended for connection between the GD402 converter and
GD40 detector.
Model
Suffix code
Description
GDW
- - - - - - - - - - - - - - - Two - core, double-shielded cable,
both ends finished with cable pins.
Length
-L
Length in meters, 500 meter maximum.
IM 11T03E01-01E
1-8
<1. Specifications>
1.2.5
Brain Terminal (Optional)
Model
Suffix code Option code
Description
BT200
- - - - - - - - - - - - - - - - - - Brain terminal *1
Printer
-N
-P
–
- - - - - - - - - Standard type (without printer)
- - - - - - - - - With printer
00
Options
*1:
- - - - - - - - - Always 00
/
The BT200 has following accessories, two communication cables, one with IC clips and another with alligator clips, handy
carrying case and five AA 1.5 V dry batteries.
n OPTIONS FOR BT200
Options
Description
Option codes
Communication cable
*1
With a 5-pin connector
(for the signal conditioner)
/C1
Intrinsically safe type
*1, *2
CSA Intrinsically safe approval
Class I, Groups A, B, C and D Temp. Code : T4
/CS1
*1:
*2:
Optional code /C1 can not be combined with /CS1.
Applicable only for Model BT200-N00.
See GS 01C00A11-00EN for “BT200” brain terminal in detail.
1.2.6
Pressure transmitter (Optional)
/EJAJ1 means TIIS certified EJX310A.
/EJAF2 means general purpose model EJX310A.
/EJAF3 means FM certified EJX310A.
/EJAF4 means CSA certified EJX310A.
See GS 01C25D01-01EN for “EJX310A” pressure transmitter in detail if a different
selection from pre-selected options seems necessary.
IM 11T03E01-01E
1-9
<1. Specifications>
1.3
External Views and Dimension
1.3.1
GD402G Converter (General purpose)
Hood
(Option code:
Unit: mm (in.)
/H)
184 (7.2)
4-M6, depth 8
220 (8.7)
80 (3.1)
20 (0.8)
144(5.7)
72 (2.8)
80 (3.1)
144 (5.7)
23
112 (4.4)
(0.9)
Cable inlet (21 mm in dia.)
Compatible with a PG13.5 cable gland
B
E
D
36
C
F
36
(1.4) (1.4)
Grounding terminal (4 mm screws)
36
38
(1.4) (1.4)
A
Cable gland
Connection
A, B
• Pressure transmitter
• Analog output
• Contact Input
C
D, E
• Detector
• Contact Output
F
• Power Supply
IM 11T03E01-01E
1-10
<1. Specifications>
1.3.2
Pipe- and Wall-Mounting Bracket (Optional)
Unit: mm (in.)
• Hardware for Pipe Mounting (Option code /U)
188 (7.4)
50
(2)
200 (7.9)
174 (6.9)
4 - M6 screws
Optional
hardware
100 (3.9)
Mounting pipe of JIS 50A (60.5 mm in outer dia.)
• Hardware for Wall Mounting (Option code /U)
13
(0.51)
4 - M6 screws
4-Ø10 holes
• Hardware for Panel Mounting (Option code /PA)
4 - M6 screws
Optional
hardware
Dimensions of panel cutout
139 +10
(5.47 +0.08
0 )
195 (7.7)
(0.9)
max. 12
(panel thickness)
100 (3.9)
139 +10
23
70 (2.8)
(5.47 +10 )
15
(0.6)
Optional
hardware
35
(1.4)
200 (7.9)
224 (8.8)
135 (5.3)
100 (3.9)
178 (7)
185 (7.3)
IM 11T03E01-01E
1.3.3
1-11
<1. Specifications>
GD402T, V, R Converter (Explosion-proof)
Unit: mm (in.)
Mounting pipe of JIS 50A nominal size
(60.5 mm (2.4) in outer dia.)
Pipe-mounting hardware
(300) (11.8)
Approx. 200 (7.9)
Approx. 212 (8.3) in dia.
Approx. 209 (8.2)
Approx. 182 (7.2)
15 (0.6)
36 (1.4)
Approx. 173 (6.8)
Approx. 242 (9.5)
140 (5.5)
(259) (10.2)
A
Side view from arrow A direction
4-M8, depth 20
Cable glands
not included
2-M8, depth 16,
one on each side
Approx. 200 (7.9)
(2.1) (1.5)
54
39 24 (0.9)
(Used to fix the pipe-mounting hardware)
A
B
C
D
E
F
Six wiring holes, G3/4 threaded
Grounding terminal
(5 mm (0.2) screw)
182 (7.2)
Ø
1
8±
19
Panel Cutout
140 (5.5)
4 holes with
10mm in dia.
Connection
Cable gland
56
56
(2.2)
(2.2)
A
• Power Supply
B, C
• Contact Output
D
• Detector
E, F
• Pressure from Transmitter,
• Analog Output • Contact Input
The screw hole for panel installation
on the converter and 20mm depth.
The bolts attached to the product are
M8 screw and 20 mm in length.
IM 11T03E01-01E
1.3.4
<1. Specifications>
1-12
GD40 Detector
l Model GD40- /EJAJ1, /EJAF2, /EJAF3, /EJAF4
approx. 443 (17.5)
Unit: mm (in.)
Detector
approx.
120 (4.7)
Pressure
transmitter
Cable gland is included
only in /EJAJ1(T)
Gas out
See Table
100 (3.9)
150 (5.9)
320 (12.6)
350 (13.8)
Gas in
See Table
EJX wiring port
See Table
350 (13.8)
GD40
Detector wiring port
See Table
300 (11.8)
70
(2.8)
4-Ø12 (0.47) holes
Option code GD40 wiring EJX wiring
/EJAJ1(T)
G1/2
G1/2
/EJAF2(T)
1/2NPT
1/2NPT
/EJAF3(T)
1/2NPT
1/2NPT
/EJAF4(T)
1/2NPT
1/2NPT
Gas out/in
Rc1/4
1/4NPT
1/4NPT
1/4NPT
IM 11T03E01-01E
1.3.5
1-13
<1. Specifications>
Detector Unit (Intrinsically Safe, Explosion-proof)
l Bracket for Pipe Mounting : GD40
Unit: mm (in.)
Approx. 191 (7.5)
Approx. 264 (10.4)
Approx. 92 (3.6)
Approx. 193 (7.6)
Grounding terminal
(3 mm (0.11) screw)
Pipe-mounting
hardware
Mounting pipe
(JIS 50 A (60.5 mm in
outer dia.) nominal size)
Wiring hole,
GD40R: G1/2
GD40V: 1/2NPT
GD40T: 1/2NPT
GD40G: 1/2NPT
Approx. 117 (4.6)
Sample gas inlet
GD40R: Rc1/4
GD40V: 1/4NPT
GD40T: 1/4NPT
GD40G: 1/4NPT
Sample gas outlet
GD40R: Rc1/4
GD40V: 1/4NPT
GD40T: 1/4NPT
GD40G: 1/4NPT
Note: Cable gland is included only in GD40R.
IM 11T03E01-01E
Blank Page
2.
2-1
<2. Installation, Wiring and Piping>
INSTALLATION, WIRING AND PIPING
The GD402 Gas Density Meter is thoroughly inspected at the factory and carefully packed to
ensure the equipment does not suffer any damage during transportation.
The package should also be handled with care when unpacking to prevent the equipment from
undergoing severe mechanical shock. After unpacking, visually check the equipment to ensure
that it is free from any damage.
Although the detector has no controls on it, it may need to be accessed for inspection or for other
reasons. Install the detector in a location as close as possible to where the gas is sampled and
where maintenance can be easily carried out. The converter has a display with controls on it;
thus, you should install it so that the keys are positioned directly in front of you when working.
Note that there must be a clearance of at least 400 mm at the back of the converter GD402G,
GD402T, GD402V and GD402R since cables are wired to the back by removing the screwed
rear cover.
2.1
Installing the Detector
The GD40 detector is an explosion-proof with intrinsically safe sensor instrument. Install the
detector in a location where the following conditions are satisfied.
2.1.1
Selecting the Location
l Explosion-proof construction
Before installing the detector in an explosion-hazardous area, ensure that the area conforms to
the explosion-proof code noted above.
l No corrosive gases
Corrosive gases are not desirable because they may damage the electrical components in the
detector.
l Slight mechanical vibration
Although the detector is vibration-resistant, vibration may loosen the connections of the external
wiring.
l No exposure to direct sunlight
Exposure to direct sunlight may raise the temperature in the detector to abnormal levels, and
should therefore be avoided. Note that such abnormal temperature levels can also result from
heat irradiated from high-temperature equipment around the detector.
l Humidity maintained between 5% and 95%RH
Avoid choosing a location that is likely to be exposed to abnormally high or low humidity over a
prolonged period. It is recommended that the detector be used at a humidity between 25% and
85 % RH.
l No exposure to rain water
Even though the detector is rainproof, whenever possible install it where it is protected from
water splashes. The reasoning for this is that the detector cover may need to be removed for
maintenance or for other reasons.
l Altitude of installation site is lower than 2,000 m.
IM 11T03E01-01E
2.1.2
2-2
<2. Installation, Wiring and Piping>
GD40T (FM Explosion-proof and Intrinsically Safe
Approval)
NON-HAZARDOUS
LOCATIONS
HAZARDOUS
LOCATIONS
475 mm (18") Max.
Sealing Fitting
Conduit
GD40T
F0201.ai
Figure 2.1
1. GD40T Gas density detector for use in hazardous (classified) locations:
• Explosion-proof for Class I, Division 1, Groups B, C and D;
Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G with Intrinsically
Safe sensor for Class I, II, III, Division 1, Groups B, C, D, E, F and G.
• Enclosure Rating:
NEMA 4X
• Temperature Code:
T5
• Ambient Temperature:
-10 to 60°C
2. Wiring
• All wiring shall comply with National Electric Code ANSI/NAPA 70 and Local Electrical
Codes.
WARNING
Seal all conduits within 475 mm (18 inches). Refer to Figure 2.1.
3. Operation
• Note a warning label worded as follows.
WARNING
Open circuit before removing cover.
• Take care not to generate mechanical spark when access to the instrument and
peripheral devices in hazardous locations.
4. Maintenance and Repair
• The instrument modification or parts replacement by other than authorized representative
of Yokogawa Electric Corporation is prohibited and will void the approval of Factory
Mutual Research Corporation.
IM 11T03E01-01E
2.1.3
2-3
<2. Installation, Wiring and Piping>
GD40V (CSA Explosion-proof and Intrinsically Safe
Approval)
NON-HAZARDOUS
LOCATIONS
HAZARDOUS
LOCATIONS
500 mm (20") Max.
Sealing Fitting
Conduit
GD40V
F0203.ai
Figure 2.2
1. GD40V Gas density detector for use in hazardous (classified) locations:
• Explosion-proof for Class I, Division 1, Groups B, C and D;
Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G with Intrinsically
Safe sensor for Class I, II, III, Division 1, Groups B, C, D, E, F and G.
• Enclosure Rating:
Type 4X
• Temperature Code:
T5
• Ambient Temperature: -10 to 60°C
2. Wiring
• All wiring shall comply with Canadian Electrical and Local Electrical codes.
• Note a warning label worded as follows.
WARNING
A seal shall be installed within 500 mm (20 inches) of the enclosure. Refer to Figure 2.2.
3. Operation
• Note a warning label worded as follows.
WARNING
• Open circuit before removing cover.
• Substitution of components may impair intrinsic safety.
• Take care not to generate mechanical spark when access to the instrument and
peripheral devices in hazardous locations.
4. Maintenance and Repair
• The instrument modification or parts replacement by other than authorized representative
of Yokogawa Electric Corporation is prohibited and will void the certification of CSA
International.
IM 11T03E01-01E
2.1.4
2-4
<2. Installation, Wiring and Piping>
Mounting the Detector
The detector is designed for pipe mounting.
Unit: mm (in.)
Approx. 191 (7.5)
Approx. 264 (10.4)
Approx. 92 (3.6)
Approx. 193 (7.6)
Wiring hole,
GD40R: G1/2
GD40V: 1/2NPT
GD40T: 1/2NPT
GD40G: 1/2NPT
Pipe-mounting
hardware
Approx. 117 (4.6)
Grounding terminal
(3 mm (0.11) screw)
Sample gas inlet
GD40R: Rc1/4
GD40V: 1/4NPT
GD40T: 1/4NPT
GD40G: 1/4NPT
Sample gas outlet
GD40R: Rc1/4
GD40V: 1/4NPT
GD40T: 1/4NPT
GD40G: 1/4NPT
Mounting pipe
(JIS 50 A (60.5 mm in
outer dia.) nominal size)
Note: Cable gland is included only in GD40R.
Figure 2.3
Vertical Mounting
IM 11T03E01-01E
2.2
<2. Installation, Wiring and Piping>
2-5
Installing the Converter
The GD402 converter comes in either the non-explosion protected Model GD402G or explosionprotected Model GD402T, V, R (explosion-protection code: Refer to 1.1.1).
Choose the Model GD402T or V or R if the converter is to be used in a hazardous area.
2.2.1
Selecting the Location
l Ease of operation
Select a location where you can easily view the readings on the display and work with the
keys. Installing the converter closer to the detector will ease your maintenance work, including
calibration.
l Explosion-proof construction
When installing the converter in an explosion-hazardous area, ensure that the area conforms to
the explosion-proof code noted above.
l No corrosive gases
Corrosive gases are not desirable because they may damage the electrical components in the
converter.
l Slight mechanical vibration
Although the converter is vibration-resistant, vibration may loosen the connections of the external
wiring.
l No exposure to direct sunlight
Exposure to direct sunlight may raise the temperature in the converter to abnormal levels, and
should therefore be avoided. Note that such abnormal temperature levels can also result from
heat irradiated from high-temperature equipment around the converter. If the converter is likely to
receive direct sunlight and the temperature in the converter will likely exceed the operating limits,
attach a hood (optional).
l Humidity maintained between 5% and 95%RH.
Avoid choosing a location that is likely to be exposed to abnormally high or low humidity over a
prolonged period. It is recommended that the converter be used at a humidity between 25% and
85%RH.
l No exposure to rainwater
Even though the converter is rainproof, whenever possible install it where it is protected from
water splashes. The reasoning for this is that the converter cover may need to be removed for
maintenance or for other reasons
l Altitude of installation site is lower than 2,000 m.
IM 11T03E01-01E
2.2.2
<2. Installation, Wiring and Piping>
2-6
GD402T (FM Explosion-proof Approval)
GD402T
Conduit
HAZARDOUS
LOCATIONS
475 mm (18") Max.
Sealing Fitting
NON-HAZARDOUS
LOCATIONS
F0202.ai
Figure 2.4
1. GD402T Gas density converter for use in hazardous (classified) locations:
• Explosion-proof for Class I, Division 1, Groups B, C and D;
Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G
• Enclosure Rating:
NEMA 4X
• Temperature Code:
T6
• Ambient Temperature: -10 to 55°C
2. Wiring
• All wiring shall comply with National Electrical Code ANSI/NFPA 70 and Local Electrical
Codes.
WARNING
Seal all conduits within 475 mm (18 inches). Refer to Figure 2.4.
3. Operation
• Note a warning label worded as follows.
WARNING
Open circuit before removing cover.
• Take care not to generate mechanical spark when access to the instrument and
peripheral devices in hazardous locations.
4. Maintenance and Repair
• The instrument modification or parts replacement by other than authorized representative
of Yokogawa Electric Corporation is prohibited and will void the approval of Factory Mutual
Research Corporation.
IM 11T03E01-01E
2.2.3
2-7
<2. Installation, Wiring and Piping>
GD402V (CSA Explosion-proof Approval)
GD402V
HAZARDOUS
LOCATIONS
Conduit
500 mm (20") Max.
Sealing Fitting
NON-HAZARDOUS
LOCATIONS
F0204.ai
Figure 2.5
1. GD402V Gas density converter for use in hazardous (classified) locations:
• Explosion-proof for Class I, Division 1, Groups B, C and D;
Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G
• Enclosure Rating:
Type 4X
• Temperature Code:
T6
• Ambient Temperature: -10 to 55°C
2. Wiring
• All wiring shall comply with Canadian Electrical Code and Local Electrical Codes.
• Seal all conduits within 500 mm of the enclosure. Refer to Figure 2.5.
• Note a warning label worded as follows.
WARNING
A seal shall be installed within 500 mm (20 inches) of the enclosure.
3. Operation
• Note a warning label worded as follows.
WARNING
Open circuit before removing cover.
• Take care not to generate mechanical spark when access to the instrument and
peripheral devices in hazardous locations.
4. Maintenance and Repair
• The instrument modification or parts replacement by other than authorized representative
of Yokogawa Electric Corporation is prohibited and will void the certification of CSA
International.
IM 11T03E01-01E
2.2.4
<2. Installation, Wiring and Piping>
2-8
Mounting the Converter
(1) Mounting the GD402G Converter
● Pipe Mounting
Bracket mounting screws
Converter
Bracket
Pipe mounting bracket
Nut (2 pcs)
U-bolt
Washers (2 pcs)
Stanchion(Ø60.5 mm pipe)
Figure 2.6
F0206.ai
Pipe Mounting
● Wall Mounting
Figure 2.3 shows how to mount the converter on a wall.
Converter
Mounting hole (3 places)
Note: The converter for wall
mounting comes with the same
mounting hardware as that of a
pipe-mounting model.
When installing the converter on
a wall, use the bracket only.
M8 bolts (not supplied)
Provide bolts of a length suitable
for the mounting holes.
Bracket
F0207.ai
Unit : mm
144
102
Center of
the converter
35
70
3-M8 screw-hole or
3-Ø10 through-holes
Dimension of cutout for wall mounting
Figure 2.7
F0208.ai
Wall Mounting
IM 11T03E01-01E
2-9
<2. Installation, Wiring and Piping>
● Panel Mounting
Figure 2.4 shows how to mount the converter on a panel.
+2
Panel
Unit : mm
139 0
Converter
Bracket
+2
139 0
Insert the converter in
the panel cutout before
attaching the brackets.
Fixing screws (2 pcs)
Dimensions of cutout for panel mounting
F0209.ai
Figure 2.8
Panel Mounting
(2) Mounting the GD402T, V, R Converter
Unit : mm (in.)
Weight : approx. 15 kg (33.1 pounds)
Pipe-mounting bracket
∅212 (8.3)
Mounting pipe of JIS 50 A nominal size (∅60.5)
200 (7.9)
209 (8.2)
182 (7.2)
173 (6.8)
36 (1.4)
242 (9.5)
140 (5.5)
A
4 - M8 depth 20
Side view from arrow A direction
Cable glands not included
54 39 24 (0.9)
(2.1) (1.5)
200 (7.9)
Grounding terminal (M5)
Figure 2.9
2 - M8 depth 16 (one on each side)
(Used to fix the pipe-mounting bracket)
A
B
C
D
E
F
56
56
(2.2) (2.2)
Mounting Explosion-proof Converter
Six wiring holes, G3/4 threaded (GD402R)
1/2 NPT threaded (GD402V, T)
F0210.ai
IM 11T03E01-01E
2.3
<2. Installation, Wiring and Piping>
2-10
Piping
CAUTION
In the case of replacement range, H2 in CO2 or Air in CO2, the output happen to be fluctuated by
pressure loading to detector. In order to reduce the effect, flowmeter is to be set upper side of
detector so that the pressure loading to the detector is going to be about atmospheric pressure.
The piping connected to the analyzer comprises a line that feeds the gas under measurement
to the detector, a line that returns (or releases to the atmosphere) the gas exhausted after
measurement, and lines that carry the zero and span calibration gases. The type of piping
chosen depends on the composition of the gas under measurement, its pressure, the amount of
dust mixed, and the response (dead time). It is advisable however to use stainless steel piping of
sizes from 6 mm and 4 mm in outer and inner diameters up to JIS 15 A (21.7mm in outer dia.).
Refer to the following instructions when connecting the analyzer piping.
l Figure 2.10, "System Configuration," shows an example of piping. Connect the lines
securely to ensure that there is no gas leakage in the system.
l Make sure the pressure of the sample gas is no greater than 0.5 MPa when measured at
the detector inlet. If the pressure is too high, use a pressure reducing valve to regulate the
pressure to a normal level. If the pressure is too low, it must be raised using a pump.
l If the gas under measurement contains dust, mist or moisture, such impurities must be
removed. Install a filter, mist separator or dehumidifier to remove these impurities from the
gas.
l In order to return the gas exhausted after measurement, the difference in pressure between
the detector inlet and the point of returning the gas must be 0.5 kPa minimum. Choose
piping with a large inner diameter to minimize pressure loss in the return gas line.
l ALWAYS install a stop valve at the point where the gas is sampled (or returned).
l The pressure transmitter for compensating pressure is designed to detect the pressure
inside the detector; install a pressure lead pipe as close to the detector as possible
(preferably, no more than 0.5 m away).
IM 11T03E01-01E
2-11
<2. Installation, Wiring and Piping>
Switching valve
Sample gas line
Filter
*1
P1
Flowmeter
*3
*2
P2
*4
Gas for zero
calibration
EJX (Explosion-proof
pressure transmitter)
*4
Gas for span
calibration
Supplied by customer.
Detector unit (EJX and GD40 detector
mounted on plate)
(Supplied by Yokogawa)
Pressure regulator
for gas cylinder
*1: P1 (Inlet pressure) <
= Max. 0.5 MPa (71 psi)
*2: P1 (Inlet pressure) - P2 (Outlet pressure) >
= 0.5 kPa (0.071 psi) (depending on the size and length of the pipe)
*3: Flowrate = 0.1 to 1 L/min
*4: The cylinder pressure must be reduced to P1 (Inle t pressure).
F0211.ai
(1) Example of Gas density and Calorie meters
Explosion-proof
pressure transmitter
P
Generator
Calibration gas
CO2
*4
H2
P2
*1
Vent
P1
Flowmeter
*2
*1 : P1 (Inlet pressure ) <
= Max 0.5885 MPa (abs.)
*2 : Flowrate = 0.1 to 1 L/min
*3 : P1 (Input pressure ) - P2 (Output pressure ) >
= 0.5 kPa
*4 : The cylinder pressure must be reduced to P1 (Inlet pressure).
Figure 2.10
*3
Detector unit
F0212.ai
(2) Example of Hydrogen purity/Replacement meter
System Configuration
IM 11T03E01-01E
<2. Installation, Wiring and Piping>
2-12
2.4 Wiring
WARNING
Danger High Voltage !
Some parts of the meter’s internal assembly have high voltages. Inadvertent contact with those
parts may result in electrical shock or injury. ALWAYS turn off the power to the meter before
removing the rear or front cover by using external circuit breaker.
This section explains how to wire the GD402 Analyzers. Note that this document is limited to the
basic system configuration only (detector, converter and pressure transmitter).
For details on the wiring of instruments used to receive analog output signals or contact output
signals, see their respective instruction manuals.
CAUTION
The GD402 gas density meter is shipped after adjusting the both of detector and converter in
pairs. When installation, confirm whether the serial number on both of converter and detector are
in pairs or not. If mismatched in pairs, converter is to be out of order. When converter or detector
supplied individually, enter the detector constants, described on inside the lid of the GD40, into
converter so that the GD402 is going to be well.
In detail, please refer to Figure 5.30, Figure 6.31 or Figure 7.28.
CAUTION
After wiring to the terminal block, do not push the remaining cables into the enclosure.
2.4.1
Wiring Procedure
(1) The types of cables wired to the GD402 Analyzers are:
(2)
• Cable wired to power supply
• Cable wired to detector input
• Cables wired to output signals (two) -- or one signal if BRAIN communication is used
• Cables wired to contact outputs (five) -- provided as necessary
• Cable wired to contact input -- provided as necessary
• Cables wired to pressure transmitter
• Cables wired to the ground
The GD402 converter has six cable inlets for external wiring. Wire the converter through
cable glands. As a rule, choose cable glands as shown in Figure 2.11.
IM 11T03E01-01E
2-13
<2. Installation, Wiring and Piping>
Unit : mm (in.)
Cable inlet (21 mm in dia.)
Compatible with a PG 13.5 cable gland
D
B
E
C
Cable gland
36
38
(1.5) (1.4)
A
F
A, B
C
D, E
F
36 36
(1.4) (1.4)
•
•
•
•
•
Connection
Pressure transmitter
Analog Output • Contact Input
Detector
Contact Output
Power Supply
Weight : Approx. 3 kg (6.6 pounds)
(including mounting bracket)
Grounding terminal (4 mm screws)
F0213.ai
54 39 24 (0.9)
(2.1) (1.5)
Grounding terminal (M5)
Unit : mm (in.)
A
B
C
D
E
F
56
56
(2.2) (2.2)
Connection
Cable gland
A
• Power Supply
B, C
• Contact Output
D
• Detector
• Pressure Transmitter
E, F
• Analog Output • Contact Input
Six wiring holes, G3/4 threaded (GD402R)
1/2 NPT threaded (GD402V, T)
F0214.ai
(3) The following explains the general wiring procedure for each type of converter.
WARNING
Cables that withstand temperatures of the least 75°C should be used for wiring.
CAUTION
The rear cover and the metal plugs at the external-wiring connections are fixed onto the
converter case with hex setscrews. Before removing the cover, be SURE to loosen the
setscrews. If you rotate the cover or a plug without loosening the setscrews, their threads may be
severed, making the cover unremovable.
When the rear cover or any of the cable glands or plugs are removed, handle them with utmost
care so that the threads are not damaged. When re-attaching it on to the converter, clean the
threads so they will not be damaged due to such foreign matters as dust.
•
•
•
Loosen the setscrews and remove the rear cover by rotating it counterclockwise.
Guide the required external-wiring cables through the cable glands to the converter.
Exercise care so as to avoid connecting them with the wrong polarities.
After wiring, securely fasten the rear cover and cable glands.
IM 11T03E01-01E
2-14
<2. Installation, Wiring and Piping>
• Loosen the setscrews in the four corners and remove the front cover.
• Remove the terminal cover. Guide the required external-wiring cables through the cable
glands to the converter. Beware of the correct polarities.
• After wiring, reinstall the terminal cover and front cover in place.
GD402 Converter
14
MAINTENANCE
Contact output
15
MAINT
+
CONT
INP -
ALM
ANLG +
OUT1
-
FAIL
ANLG +
OUT2
-
16
ALARM
Contact output
17
18
FAIL
Contact output
19
1
2
3
4
Contact input
Isolated 4-20 mA Output
with BRAIN Communication
5
6
Isolated 4-20 mA Output
20
SPAN/FUNC
22
SELECT GAS
Contact output
23
24
Power supply
100 to 240 V AC or 24 V DC
25
*3 26
ZERO/SEL GAS
L or +
SNSR +
PWR
SNSR +
INP
-
N or G
Case grounding
terminal
DET +
INP
SHIELD
7
8
Pressure Transmitter
-
9
10
11
+
Case grounding
terminal
GD40 Detector
12
13
+
21
-
FUNCTION
Contact output
Case grounding
terminal
Intrinsic safety
Grounding *1
Class A grounding
(grounding resistance
10 Ω or less) *2
Cable List
Terminal
Indication
Shield
Requirement
Requirement
MAINTENANCE MAINT
Contact output
Unshielded
ALARM
Contact output
ALM
Unshielded
FAIL
Contact output
FAIL
Unshielded
FUNCTION
Contact output
SPAN
Unshielded
SELECT GAS
Contact output
ZERO
Unshielded
Contact input
CONT IN
UnShielded
Analog output1
ANLG OUT1
Shielded
Analog output2
ANLG OUT2
Shielded
Total resistance should not exceed 50 Ω.
Shield should be grounded at one end only.
Maximum load resistance including wire resistance is 600Ω.
When BRAIN communication is used, it is 250 to 550 Ω.
Pressure
SNSR PWR
transmitter input SNSR INP
Shielded
Total resistance should not exceed 50 Ω.
Shield should be grounded at one end only.
Total resistance should not exceed 50 Ω.
Connect shield to SHIELD terminal on converter.
Detector input
DET INP
SHIELD
Shielded
Supply
L, N, G
Unshielded
*1 Intrinsic safety grounding
GD402V, GD40V; All wiring should comply with Canadian Electrical Code and Local Electrical Codes.
GD402T, GD40T; All wiring should comply with National Electrical Code and ANSI/NFPA 70 and
Local Electrical Codes.
*2 When select GD402T/V/R.
*3 Terminal 26 is connected to the case-grounding terminal.
Figure 2.11
Terminals on the Converter
IM 11T03E01-01E
2-15
<2. Installation, Wiring and Piping>
Electrical Noise Protection
If a malfunction due to noise occurs, strengthen measures against noise.
For example, ground the detector body or use a double-shielded cable. If a double-shielded
cable is used, ground shields of each conductor at one end. Ground one end of the outer shield
on the detector side to the detector case and connect the other end on the converter side to
terminal 13. See the User’s Manual for more instructions on cable installation.
Contact Input Function of the Hydrogen Purity Meter
For hydrogen purity meter, the contact input is used for range selection.
Open: Concentration measurement for Air in CO2
Close: Concentration measurement for H2 in CO2.
Contact Output Specifications
Indication
MAINT
ALM
Specification
Contact Type:
Voltage free, dry contact (mechanical relay contact output)
Contact rating:
250 V AC 3 A or 30 V DC 3 A
Contact arrangement: NO/NC, selectable
FAIL
Contact Type:
Voltage free, dry contact (mechanical relay contact output)
Contact rating:
250 V AC 3 A or 30 V DC 3 A
Contact arrangement: NC, fixed
SPAN/FUNC
Function contact:
ZERO/SEL GAS
Use distinguish between the Hydrogen purity meter and
the Replacement meter.
Select gas contact: Use distinguish measuring ranges in the Replacement meter.
Contact Type:
Voltage free, dry contact (mechanical relay contact output)
Contact rating:
250 V AC 3 A or 30 V DC 3 A
Contact arrangement: NO/NC, selectable
IM 11T03E01-01E
2.4.2
<2. Installation, Wiring and Piping>
2-16
Cables Wired to Power Supply
The GD402G bears the CE marking. When using the meter in a place where the CE marking is
obligatory, or when performance meeting the CE marking requirements is needed, the following
wiring is required. (Note that the following wiring is not required when the power supply is 24 V
DC.)
● Install an external switch or circuit breaker to the power supply of the converter.
● Use an external switch or circuit breaker that is rated at 5 A and conforms to IEC 947-1 or
IEC 947-3.
● Install the external switch or circuit breaker in the same room as the converter is installed.
● The external switch or circuit breaker should be installed in a location that allows operator
access and should be marked a power supply switch to the converter.
CAUTION
The GD402R, T, V converter contains a power switch and a fuse. The switch cannot be operated
however, if the converter is installed in an explosion-hazardous area; it must be left placed in the
ON position. When Power ON/OFF, use external circuit breaker
Wiring is for supplying power supply that meets specifications to the GD402 converter.
Use a 3-conductor cable with a size of 1.25 to 2.5 mm2 and an outside diameter of 8 to 16 mm
for 100-240 V AC type and a 2-conductor cable for 24 V DC type.
Wire the cable as instructed below:
(1) Strip the sheath 7 mm at the cable end to be connected to the converter.
(2) Connect the cable to terminals L, N, and G or terminals + and - on the converter.
Loosen the terminal screw, insert the stripped part of the wire into the terminal hole,and fix
the cable by tightening the screw. (adequate tightening torque: 0.4 N·m)
GD402 Converter
24 V DC type
GD402 Converter
100-240 V AC type
Case-grounding
terminal
SUPPLY
LN G
Case-grounding
terminal
External circuit breaker
Rating: 5 A
Use IEC947-1 or IEC947-3 compliant product.
SUPPLY
+ -G
24 V DC
100 to 240 V AC
50/60 Hz
Figure 2.12
Cables Wired to Power Supply
IM 11T03E01-01E
2.4.3
<2. Installation, Wiring and Piping>
2-17
Cables Wired to Outputs
These cables are used to transmit 4-20 mA DC signals and carry out BRAIN communication.
Use shielded cables of 8 to 16 mm in finished outer diameter and 0.75 mm2 minimum in
thickness (or a two-core shielded cable for single output).
Wire the cables as instructed below:
(1) Use cables that are 0.75 to 2.5 mm2 thick. Treat the cable ends by stripping the core wires
back 7 mm (see Figure 2.14).
(2) Wire the cable for output 1 to terminals 3 and 4 on the block and the cable for output 2 to
terminals 5 and 6 on the block. Beware of the correct polarities. Note that only terminals 3 and
4 (output 1) are effective for BRAIN communication.
Loosen the terminal screws, insert the stripped ends of the core wires into the terminals, and
fasten the screws to fix the cables.
(3) Ground the shielding wire at a given terminal on the converter; do not ground the other end of
the cable. (The appropriate tightening torque for the internal wiring terminals is 0.4 N·m.)
7 mm
F0217.ai
Figure 2.13
Examples of Treatment on Cable End Wired to the Converter
IM 11T03E01-01E
2.4.4
<2. Installation, Wiring and Piping>
2-18
Cables Wired to Contact I/Os
The contact I/Os of the GD402 converter comprise the start-of-calibration contact (input), the
FAIL, ALARM and MAINTENANCE contacts, and the contacts for operating the solenoid valves
for span and zero calibrations. Choose the type of cables for a group of calibration-purpose
contacts and for a group of other contacts separately.
For wiring to the calibration contact output, use a cable of 8 to 16 mm in finished outer diameter
and 1.25 mm2 minimum in thickness. For wiring to other contacts, use a cable of 8 to 16 mm in
finished outer diameter and 0.13 to 1.25 mm2 in thickness, while choosing the number of core
wires depending on the number of contacts used.
Since the contact outputs are voltage-free, driving such devices as alarm lamps requires a
separate power supply. For the contact input (start of calibration), provide a no-voltage signal.
(Do not apply an external voltage.)
Wire the cables as instructed below:
(1) Use cables 0.13 to 1.25 mm2 thick. Treat the cable ends by stripping back the core wires to 7
mm from the end (see Figure 2.14).
(2) Wire the cables to their respective given terminals on the block, being careful not to mistake
one terminal for another. For terminal numbers assigned to the respective contacts I/Os, see
Figure 2.12.
Loosen the terminal screws, insert the stripped ends of the core wires into the terminals, and
fasten the screws to fix the cables. (The appropriate tightening torque is 0.4 N·m)
Terminal
Tightening screw
Cable inlet
Cable
Figure 2.14
F0218.ai
Connection between terminal and cable
IM 11T03E01-01E
2.4.5
<2. Installation, Wiring and Piping>
2-19
Cable Wired to GD40 Detector
The cable wired to the GD40 detector comes from the GD402 converter. Use a shielded cable
of 10 to 13.5 mm in finished outer diameter that suits the dedicated cable gland. Wire the cable
through the cable gland, being careful not to connect the cable with the wrong polarities. Attach
M4 crimp terminals to the ends of the core wires (see Figure 2.14). Connect the shielding wire to
terminal 13 on the converter (see Figure 2.12). When crimping the terminals, use tools that fit the
terminals.
If a malfunction occurs and it is assumed to be due to noise, use a double-shielded cable (for
example, the cable with Yokogawa’s Model GDW-L ). Connect the outer shielding wire
to terminal 13 on the converter and the other end of the wire to the grounding terminal on the
detector. Connect the inner shielding wires to the grounding terminal on the converter (see Figure
2.17).
Class A
grounding terminal
Terminals for
wiring to converter
F0219.ai
Figure 2.15
Configuration of Detector Terminals
GD40T, V, R
Detector
GD402G, T, V, R
Converter
Class A
grounding
terminal
11 12 13
Class A
grounding
Inner shielding
wires
Outer shielding wire
Do not connect the inner
shielding wires here.
Outer shielding
wire
F0220.ai
Figure 2.16
Connection Between Detector and Converter When Using a Double-Shielded Cable
IM 11T03E01-01E
2.4.6
<2. Installation, Wiring and Piping>
2-20
Wiring to Model EJX310A pressure transmitter
When Wiring to model EJX310A pressure transmitter, connect by using cables from GD420o
converter. Wire the cables through cable glands. Be sure not to misconnect polar. For finished
external diameter, refer to Figure 2.11, "Terminals on the Converter."
The size of the terminal screw threads is M4. Use a crimp terminal suited for your cable terminal.
EJX310A
Pressure transmitter
GD402G, T, V, R
Converter
7 8 9 10
AL
CHK
SUP.
+
-
Figure 2.17
Connection Between Detector and Pressure Transmitter
For wiring and grounding of the pressure transmitter, refer to the user’s manual attached to the
pressure transmitter.
2.4.7
Cables Wired to the Ground
When using the GD40R detector in an area requiring explosion protection, BE SURE to ground it
as per the Class A Grounding Standard using a conductor of at least 2 mm2 in nominal thickness.
Also BE SURE to ground the GD402 converter as per the Class D Grounding Standard.
In the case of using GD40T detector, Intrinsically Safe Grounding shall comply with National
Electrical Code ANSI / NFPA 70 and Local Electrical Codes.
In the case of using GD40V detector, Intrinsically Safe Grounding shall comply with Canadian
Electrical Code and Local Electrical Codes.
Use No. 26 internal wiring terminals or the grounding terminals of the converter. (The appropriate
tightening torque for the internal wiring terminals is 0.4 N·m). Case-grounding terminal can be
used as protective ground terminal. When using the detector’s grounding terminals, attach M4
crimp terminals to the ends of the grounding cables. When crimping the terminals, use tools that
fit the terminals.
Joining advice of Converter
case ground terminal
Terminal
Converter case
Ground wiring
crimp terminal
Lock washer
Ground terminal
24 25 26
L
N
G
Earth cable
Earthing of enclosure wiring terminal
Figure 2.18
Earthing of internal wiring terminal
Power and Ground Wiring
IM 11T03E01-01E
<3. Operation>
3-1
3. OPERATION
This chapter explains how to operate the GD402 Analyzers.
3.1
Preparation for Operation
WARNING
• DO NOT supply power to the analyzer when inspecting it.
• Leave the power switch which is built in the converter, in the ON position.
• DO NOT open the terminal box cover with the power turned on, if the analyzer is installed in
an explosion-hazardous area.
• When using the converter under high ambient temperatures, be careful as the metal parts of
the converter will also be hot.
3.1.1
Inspecting Installation, Piping and Wiring
Workmanship
Before turning the power on, inspect workmanship with regards to installation, piping and wiring
as instructed below:
l Examine how the GD402 analyzer and its peripheral equipment are installed.
l Check the sample-gas line, which feeds the GD40 detector, for any gas leakage.
l Inspect the system to ensure that the GD402 analyzer and its peripheral equipment are
wired correctly. After inspection, shut the terminal box cover securely.
l Place the power switch built in the GD402T, GD402V, GD402R converter (explosionprotected) in the ON position.
IM 11T03E01-01E
3.1.2
<3. Operation>
3-2
Supplying Power
First make sure that when the GD402 analyzer is turned on, no control instruments connected to
the analyzer will come into operation by the signals transmitted from the analyzer. Next, turn on
the external power switch to start the analyzer. At that point, make sure the internal power switch
is in the ON position if the converter is an explosion-protected model.
After power-on, the analyzer will begin to operate in the measurement mode.
[Major Items Shown in Measurement Mode]
l Measured Values
In the Message selection under the DISP mode, choose the item you want to show.
Item
No. of Digits and Position
of Decimal Point
Message
Physical density
XX.XXXX
KG/M3
Physical density
X.XXXXX
LB/FT3
Compensated density
X.XXXX
KG/M3
Compensated density
X.XXXXX
LB/FT3
Calorific value
XXX.XXX
MJ/M3
Calorific value
X.XXXX
KBTU
Specific gravity
X.XXXX
SP GR
Molecular weight
XXX.XX
MOL
Concentration
XXX.X
VOL%
l Types of Messages
• Measurement units
• Error number (shown if an error occurs)
• Alarm number (shown if an alarm occurs)
l HOLD and FAIL Indications (shown regardless of whether the analyzer is in the
measurement mode or not)
• HOLD appears when the analyzer is in a hold state.
• FAIL appears if an error occurs.
IM 11T03E01-01E
3.1.3
3-3
<3. Operation>
Display on Operation Panel and Operation Keys
Figures 3.1 and 3.2 illustrate the operation panels of the Model GD402G and GD402T, GD402V,
GD402R converters, respectively.
The operation panel has a display and operation keys. Use the seven operation keys on the
panel to set operation parameters, and so on.
State Display
Operation level
display panel
Data Display
Message Display
E X A GD402
MODE
DENSITY
DENSITY (COMP.)
SPEC GRAVITY
CALORIC VALUE
MOL.WEIGHT
VOL.% CONC.
MEASURE
SEL GAS RANGE
DISPLAY
SEMI CAL.
MANUAL CAL.
VALVE
Key operation Display
Operation Key
【YES】: When the blinking "Key Operation
Display" is YES, the YES key is pushed.
【NO】: When the blinking “Key operation
Display" is NO, the NO key is pushed.
【MODE】: When switching from the measurement mode to the operation mode,
the mode key is pushed. When
switching from modes other than
the measurement to the measurement
mode, the mode key is pushed.
【
】: This switch is used to select the input
data the digit.
【
】: This switch selects the input data value.
【ENT】: When the input data is entered,
this switch is pushed
Figure 3.1
MODE
NO
YES
RANGE
CAL DATA
ALARM
SERVICE
contacts
MAINT
ALARM
ENT
*
Setting level / Service
level display panel
Setting level / Service
level selection switch
CAL/SEL
FAIL
MARKINGS
WITHIN
ENCLOSURE
F0301.ai
Functional Description of GD402G Converter (Cover Open)
State Display
Data Display
Message Display
Operation level
display panel
Key operation Display
UNIT /MODE
kg/m3
SP 3R
MJ/m3
MOL WT
vol% conc
MEASURE
DISPLAY
SEMI CAL
MANUAL CAL
VALVE
Operation Key
【YES】: When the blinking
"Key Operation Display" is YES,
the YES key is pushed.
【NO】: When the blinking “Key operation
Display" is NO, the NO key is pushed.
【MODE】: When switching from the measurement mode to the operation mode,
the mode key is pushed. When
switching from modes other than
the measurement to the measurement
mode, the mode key is pushed.
【
】: This switch is used to select the input
data the digit.
【
】: This switch selects the input data value.
【ENT】: When the input data is entered,
this switch is pushed
Figure 3.2
DENS
SUB MENU
kg/m3 DENS COMP
CONTACTS
YES
MAINT
NO
ALARM
CAL/SEL
MODE
FAIL
RANGE
CAL DATA
ALARM
SERVICE
Setting level / Service
level display panel
EXA GD402
*
Setting level / Service
level selection switch
ENT
F0302.ai
Functional Description of GD402T, GD402V, GD402R Converter (Cover Open)
IM 11T03E01-01E
3.1.4
3-4
<3. Operation>
Basic Key Operation
The basics of key operation are the selection between modes/levels, the selection and execution
of the function/action, and entry of data values.
1. Selecting Between Modes at the Operation Level (See Chapter 5, 6 or 7 for details.)
l You can enter the measurement mode by pressing the [MODE] key while the analyzer is in
a mode other than the measurement mode.
When switching between modes, note the following two points:
(1) Press the [MODE] key once. The message field first shows ”DISP” in Figure 3.3-(1)
(then password is not selected in first condition). And then changes to look like one shown in
Figure 3.3-(2).
(1)
When password is selected, the message field
<*PASSW>, prompting you to enter a passwords.
Type [XXX] and press the [ENT] key. The
message field changes to look like one shown left.
Figure 3.3
(2)
In the display shown left, the [YES] and [NO]
fields blink. Pressing the [YES] key will change
the message to [KG/M3], indicating the analyzer
is now in the density mode.
F0303.ai
Examples of Display at Operation Level (In case of the Density Meter)
(2) In the display shown in Figure 3.3-(1), each press of the [NO] key changes the message
field, enabling you to choose a different mode by its indication.
2. Selecting Between Modes at the Setting Level (See Chapter 5, 6 or 7 for details.)
(1) In the measurement mode, press the [*] key (to select between settings).
(1)
When password is selected, the message field shows
<*PASSW>, prompting you to enter a passwords.
Type [XXX] and press the [ENT] key. The
message field changes to look like one shown left.
(2)
In the display shown left, the [YES] and [NO]
fields blink. Pressing the [YES] key will change
the message to read [OUT1] , indicating the
analyzer is now in the analog output setting mode.
F0304.ai
Figure 3.4
Examples of Display at Setting Level
(2) In the display shown in 3.4-(1), each press of the [NO] key changes the message field,
enabling you to choose a different mode by its indication.
IM 11T03E01-01E
3-5
<3. Operation>
3. Entering Data Values
NOTE
The data you have entered keep in memory even if you turn off the power. If the data you have
entered are provisional, reenter the normal data.
The following explains how to make changes to existing data entries.
In the display shown in Figure 3.4-(2), pressing the [YES] key causes the message field to default
to [*DENS]. Pressing the [YES] key once again changes the message to [*Z_DNS], the range of
lower limit setting, and makes the indication of the lower limit blink.
Assume the existing data entry of the range's lower limit is 1.2927. To change the value to
2.0535, follow the instructions given below:
(1) Press the [>] key so “1“ starts blinking.
>
(2) Press the [ ] key until the blinking "1" changes to ''2".
>
(3) Press the [>] key until the next digit begins blinking. Then, press the [ ] key until the
blinking ”2” digit changes to “0".
>
(4) Press the [>] key until the next digit begins blinking. Then, press the [ ] key until the blinking
digit changes to ''5”.
>
(5) Press the [>] key until the next digit begins blinking. Then, press the [ ] key until the blinking
digit changes to ''3”.
>
(6) Press the [>] key until the next digit begins blinking. Then, press the [ ] key until the blinking
digit changes to ''5”.
(7) Press the [ENT] key. This confirms the value “2.0535".
(1)
In the display shown in Figure 3.4 (2), pressing the
[YES] key will change the display to look like this.
(2)
Pressing the [ENT] key enters the value shown
in the data field.
F0305.ai
Figure 3.5
Display During Data Setting
IM 11T03E01-01E
3-6
<3. Operation>
4. Entering a Password
If you need, a password can be entered to proceed into the operation, setting or service level.
When you attempt to enter that level, a password-entry message (password prompt) like one
shown in Figure 3.6-(1) appears. The required password is “XXX”. (See section “4.2 Setting
Lists”.)
(1)
(2)
When password is selected, pressing the [MODE]
key, then the [*] key, changes the display to
look like this.
Type the password (XXX) and then press the
[ENT] key. If you type a wrong password,
you cannot enter the level.
Figure 3.6
3.1.5
F0306.ai
Password Prompt and the Reading of a Password Typed
Checking the Setting Parameters
Set up the relevant parameters, display unit, analog output and others, to meet individual
operating requirements. The GD402 gas density meter is designed to change into density,
calorific meter and hydrogen purity meter according to application. (Select “50” in Service code,
and change parameter you need.)
3.1.6
Calibrating the Analyzer for Correct Readings
For details on the calibration of the analyzer for correct readings, see Chapter 8, 9 or 10.
IM 11T03E01-01E
3.1.7
<3. Operation>
3-7
Checking the Analyzer for Performance
When the required parameters have been completely set, bring into operation all equipment
composing the measurement loop. Keep this equipment operating for a while. After making sure
there is no problem with the equipment, go into normal operation.
For reference, the following summarizes the setting functions (related to signals) of the GD402
analyzer that are used to obtain the optimum operating conditions.
[Functions Related to the Output Signal]
The output signal can be held at a value immediately before calibration or at a desired value
during calibration or at the setting/service level. This is to prevent the output signal from adversely
affecting the operation of the equipment configuring the measurement loop (default setting: the
output signal is held at the value immediately before).
[Functions Related to the High/Low Limit Alarm Contact Output]
Select the parameters for which the alarms are raised, and then set their respective high/low limit
set-points. The analyzer transmits an alarm contact signal if any of these setpoints is exceeded.
(Define the voltage-free contacts and the state of contacts at the service level.)
[Functions Related to the FAIL Contact Signal]
If a FAIL state occurs, an appropriate message appears on the LCD display. For details on these
functions, see Chapter 11.
[Contact output of Hydrogen purity / Replacement meter]
According to switching of Replacement meter, the status of contact and display (LED) is changed
as following. (The status of contact output are to be set in Service level, CODE 05.)
• Hydrogen purity meter
RANGE
FUNC
SEL GAS
LED
Replacement
Meter
H2 in CO2
AIR in CO2
—
ACTIVE
NORMAL
ON
OFF
Hydrogen Purity/
Replacement Meter
H2 in AIR
H2 in CO2
AIR in CO2
NORMAL
ACTIVE
ACTIVE
NORMAL
ACTIVE
NORMAL
OFF
ON
OFF
FUNC: Identification signal of Hydrogen purity/Replacement Meter. (contact)
SEL GAS: Range identification signal of Replacement Meter. (contact)
[The status of contact output]
Power Supply
OFF
Power Supply
ON
FAIL
SEL
FUNC
MAINT
ALM
OPEN
CLOSE
CLOSE
CLOSE
CLOSE
NORMAL: CLOSE
The status of contact are to be set in Service level, CODE 05.
FAILURE: OPEN
IM 11T03E01-01E
3.2
3-8
<3. Operation>
Normal Operation
In normal operation, there is no need for working with the GD402 analyzer except when
calibrating it. Unless there is any failure found, carry out maintenance, inspection, and so on at
the same time the analyzer is calibrated. The analyzer is designed to prohibit you from calibrating
it or setting each parameter unless you have entered the password (XXX).
(See section “4.2 Setting Lists”.)
3.2.1
Starting Operation
Follow the procedure below to start operation.
Turn on the external power switch.
Set the operating parameters.
Calibrate the analyzer for correct zero and span setpoints.
Completing these steps return to the mode.
F0307.ai
The following details the operations required in this procedure.
(1) The analyzer, when turned on, indicates the density value. See Figure 5.2, 6.3 or 7.3 for the
operations used to change to the required unit of density.
(2) Set the relevant operating parameters by checking and modifying their default values in
order to meet individual operating requirements.
(3) Supply the zero and span gases to check readings. If the readings contain errors, the
analyzer must be calibrated. For details on calibration, see Chapter 8, 9 or 10.
(4) When calibration is complete, the analyzer returns to the mode.
3.2.2
Corrective Actions in Case of Failure
If the GD402 analyzer senses a failure, it sends out a FAIL contact output signal through
terminals 18 and 19. In such a case, the output signal is held at either the value held immediately
before or at the preset value for each setting.
The message field indicates the type of error by the error code. Should a failure occur, check the
nature of the failure and take corrective action without delay. Table 3.1 summarizes the numeric
codes of errors that may occur in the measurement mode. For details on each type of error, see
Table 11.2 on page 11-4.
Table 3.1
Error No.
Errors that May Occur in the Measurement Mode
Nature of Failure
Corrective Actions
Err. 01
Sensor oscillation shutdown
Reset the power and contact service personnel.
Err. 02
Faulty oscillation frequency of sensor
Reset the power and contact service personnel.
Err. 03
Sensor’s failure to detect temperature
Contact service personnel.
Err. 04
Failure in A/D converter
Contact service personnel.
Err. 05
Failure in memory
Contact service personnel.
IM 11T03E01-01E
3.2.3
<3. Operation>
3-9
Inspection and Maintenance
Due to the operating principle (polymodal self-excited oscillation circuitry), the sensor has
only the slightest marginal drift, which is, in fact, generated by the detector itself or a result of
contamination from dust, etc. Therefore, the GD402 analyzer basically does not have to undergo
any inspection or maintenance. Periodical inspection and maintenance are recommended
however, to ensure optimum performance. For details on inspection and maintenance, see
Chapter 11.
3.3
Shutdown and Restart
3.3.1
Measures for Shutdown
Data set in the converter are retained even when the power is turned off. If the system needs to
be shut down over a prolonged period, turn off the power. Any foreign matter that accumulates
and hardens on the detector or any corrosive substance that condenses on the detector can
be the cause of problems. Fully purge the sensor with fresh air (instrument air). Cleaning of the
sensor is also recommended where it is appropriate; consult Yokogawa.
3.3.2
Measures for Restarting
Visually check all component equipment of the analyzer before you resume measurement after a
long-term shutdown of the analyzer. Re-confirm the wiring and piping have tight connections .
IM 11T03E01-01E
Blank Page
4-1
<4. Functions>
4. FUNCTIONS
When using the GD402 analyzer, set data and select functions according to the measuring
conditions. This chapter describes the structures of function.
4.1
Summary of Setting Operations
4.1.1
Measurement, Operation, Setting and Service
Parameters can be set by selecting the appropriate mode. These modes are classified into four
levels: measurement, operation, setting, and service.
[Measurement Level]
This level is used to view various measured values.
[Operation Level]
This is basically a level used to select between items to be viewed in the measurement mode
and a parameter to be measured, and to perform operations relating to routine inspection and/or
maintenance such as calibration.
[Setting Level]
At this level, there are modes for setting data related to analog outputs, calibration data and
alarm contact outputs.
[Service Level]
The GD402 analyzer has numerous functions. At this level, there are modes for selecting the
functions necessary for system operation.
4.1.2
Key Operations
Keys can be operated in an "interactive" manner. Operate the keys according to the information
in the data or message field, information pointed to by the pointer (mode indication), and/or
information in the key operation indicators. For basic key operations, see Subsection 3.1.4.
[Interactions]
l Flashing pointer
This inquires whether you want to go to the mode being pointed at or move the pointer to the
next mode. When the system indicates the modes for the setting or service level, the message
is preceded by an asterisk (*). When you enter one of the modes, the pointer stops flashing and
remains continuously lit.
l Flashing key operation indicators
This prompts you to choose an indicator from the indicators flashing on the display and press
the corresponding key.
l Flashing numerals in data field
This asks whether you want to change the flashing numeral or move to the next digit and have
it flash. Press the appropriate key. If neither of these actions is required, press the [ENT] key.
[Aborting the Setting Operation]
Press the [MODE] key. Normally, the [MODE] key is used to move from the measurement
() mode to a mode at the operation level. It is also used to return to the
measurement mode from other modes.
IM 11T03E01-01E
4.1.3
<4. Functions>
4-2
Points to Be Noted When Making Settings
(1) Password
Password is not selected in first condition. When password is necessary, refer to Section “4.2
Setting Lists” and selected password what you need. If password is selected, you are not allowed
to access any mode unless you enter the given password (XXX).
A password prompt appears when:
l at the operation level, the [MODE] key is pressed during the measurement mode, or
l at the setting level, the [*] key is pressed during the measurement mode.
l at the service level, the [YES] key is pressed with [*SERV] displayed.
(2) Related Parameters
If any change has been made to a particular parameter, check its relationship with parameters
set in other modes so that there will not be any inconsistency.
[Parameters to Check When the Analog Output Settings Are Changed]
a. High/low limit alarm set-points and contact outputs
b. Fixed value for held output signal
[Parameters to Check When High/Low limit Alarm Set-points Are Changed]
a. Alarm actions (low limit alarm or high limit alarm operation)
b. Analog output settings
4.2
Setting Lists
The GD402 Gas Density meter comes in 3 different models (and model codes) to meet several
measurement conditions: (1) Density meter, (2) Calorie meter, and (3) Hydrogen purity meter.
The values measured by the different models are as follows.
(1) Density meter : Density, compensated density, specific gravity, molecular weight,
concentration and calorie.
(2) Calorie meter : Density, compensated density, and calorie
(3) Hydrogen purity meter : Density, compensated density, H2 in Air, Air in CO2, and H2 in
CO2.
The following describes the conditions required when using the GD402 Gas Density meter.
After following the procedure below, set a model code and select a password.
(1) Select a model code for each service in “Selecting the instruments” function of each service
code. For example, if you select model 1, you can use the analyzer as a calorie meter. (See
Figure 5.29, Figure 6.30 or Figure 7.27.)
code 0 : Density meter
1 : Calorie meter
2 : Hydrogen purity meter
(2) Next, set a password, if you need. (See Figure 5.27, Figure 6.28 or Figure 7.25.)
(3) Other parameter settings can now be set.
Set parameters according to the selected model code and password.
(1) Density meter: See Chapter 5. , (2) Calorie meter: See Chapter 6. , (3) Hydrogen
purity meter: See Chapter 7.
IM 11T03E01-01E
5.
5-1
<5. DENSITY / PARAMETER SETTING>
DENSITY / PARAMETER SETTING
CAUTION
Password is not selected in first condition. When password is necessary, refer to Section “4.2
Setting Lists”.
This chapter describes how to set parameters of Density meter.
If select Calorie meter, see Chapter 6.
If select Hydrogen purity meter, see Chapter 7.
Conversion Table
1 lb/ft3
1 kBTU/ft3
= 16.01847 kg/m3
= 37.25901232 MJ/m3
X degF
= (X-32)/1.8
1 psi
= 6.894756794 kPa
Density of Air = 1.2928 kg/m3
= 0.08070684 lb/ft3
IM 11T03E01-01E
5.1
5-2
<5. DENSITY / PARAMETER SETTING>
Setting Parameters
Subsections 5.1.1 to 5.1.4 show the setting parameters for each level.
5.1.1
Setting Parameters at Measurement Level
When turned on, the analyzer starts up in the measurement mode ().
Table 5.1
Setting Parameters at Measurement Level
Mode/Setting parameter
Measurement mode
Indication of measurement value
Physical density
Compensated density
Specific gravity
Calorific value
Molecular weight
Concentration
Indication of message
Unit
Error symbol and number
Alarm mark and number
Indication of HOLD and FAIL
HOLD
FAIL
DISP mode
Physical density
Compensated density
Specific gravity
Calorific value
Molecular weight
Concentration
Sample gas temperature
Sample gas pressure
Analog output 1 (%)
Analog output 2 (%)
Display
Data to be Set
(or Conditions)
Remarks
See Figure 5.2 for key operation.
KG/M3
LB/FT3
KG/M3
LB/FT3
SP GR
MJ/M3
KBTU
MOL
VOL%
Set the unit in CODE 21.
Set the unit in CODE 21.
Set the unit in CODE 22.
The unit is the one set in the service level.
Appears if an error occurs.
Appears if an alarm occurs.
Appears even during other modes.
Appears during a HOLD state.
Appears if a failure occurs.
ERR.NO
ALM.NO
HOLD
FAIL
DISP
KG/M3
LB/FT3
KG/M3
LB/FT3
SP GR
MJ/M3
KBTU
MOL
VOL%
ºC
ºF
KPA
MPA
PSI
MA1%
MA2%
Set the unit in CODE 21.
Set the unit in CODE 21.
See Figure 5.2 for key operation.
Selection of parameters to be measured in
the measurement mode.
Set the unit in CODE 22.
Set the unit in CODE 23.
Read-only
Set the unit in CODE 20.
Read-only
Read-only
Read-only
IM 11T03E01-01E
5.1.2
5-3
<5. DENSITY / PARAMETER SETTING>
Setting Parameters at Operation Level
Press the [MODE] key and enter the password (XXX) to gain access to this level (see Figure 5.1
for key operation).
Table 5.2
Setting Parameters at Operation Level
Mode/Setting parameter
Calibration mode
Semi-automatic calibration
Manual calibration
Setting of zero-point density
Setting of span-point density
Zero/span-gas valve operation
Zero-gas valve operation
Span-gas valve operation
Sample-gas valve operation
Display
SEM.CAL
MAN.CAL
Z_DNS
S_DNS
VALVE
V_ZERO
V_SPAN
Data to be Set
(or Conditions)
One-touch calibration
Remarks
See Figure 8.3 for key operation.
See Figure 8.4 for key operation.
No setting is required for See Figure 8.1 for key operation.
manual calibration
Control of zero-gas valve
Control of span-gas valve
Switching between
sample and calibration
gases
IM 11T03E01-01E
5.1.3
<5. DENSITY / PARAMETER SETTING>
5-4
Setting Parameters at Setting Level
Press the [*] key and enter the password (XXX) to gain access to this level (see Figure 5.4 for key
operation).
Table 5.3
Setting Parameters at Setting Level
Mode/Setting parameter
Display
Analog output setting
Output 1
Set a parameter to be output
*RANGE
*OUT1
*DENS
*C_DNS
*SP_GR
*CALRY
*MOL
*CONCT
*TEMP
*PRESS
Set the zero and span points
Physical density
Compensated density
Specific gravity
Calorific value
Molecular weight
Concentration
Output 2 (same as output 1)
Data to be Set
(or Conditions)
Remarks
See Figure 5.4 for key operation.
Physical density
Compensated density
Specific gravity
Calorific value
Molecular weight
Concentration
Gas temperature
Gas pressure
No setting is required if the parameter
is temperature or pressure.
*Z_DNS,*S_DNS 00.0000 to 60.0000 kg/
m3
*Z_CP.D,*S_CP.D 0.00000 to 4.00000 lb/ft3
0.0000 to 6.0000 kg/m3
*Z_SPC,*S_SPC 0.00000 to 0.40000 lb/ft3
*Z_CAL,*S_CAL 0.0000 to 5.0000
000.000 to 133.000 MJ/
*Z_MOL,*S_MOL m3
*Z_CON,*S_CON 0.0000 to 3.5000 kBTU/
ft3
*OUT2
000.00 to 140.00
000.0 to 100.0
Setting calibration data
Zero-point density
*CAL.DT
*Z_DNS
Span-point density
*S_DNS
Output hold value during
calibration
*C_HLD
Setting of preset value
(Only if the preceding
parameter is set to 2 (preset
value))
*PR.SET
0.0000 to 6.0000 kg/m3
0.00000 to 0.40000 lb/ft3
0.0000 to 6.0000 kg/m3
0.00000 to 0.40000 lb/ft3
· Disable: 0
· Enable (value
immediately before) : 1
· Enable (preset value) : 2
-10.0 to 110.0
See Figure 8.2 for key operation.
*ALARM
Alarm-point setting
See Figure 5.5 for key operation.
Select a parameter for which an
alarm is set
*DENS
Physical density
*C_DNS
Compensated density
*SP_GR
Specific gravity
*CALRY
Calorific value
*MOL
Molecular weight
*CONCT
Concentration
Set the high/low limits
Physical density
*L_DNS,*H_DNS 00.0000 to 60.0000 kg/m3
0.00000 to 4.00000 lb/ft3
Compensated density
*L_CP.D,*H_CP.D 0.0000 to 6.0000 kg/m3
0.00000 to 0.400000 lb/ft3
Specific gravity
*L_SPC,*H_SPC 0.0000 to 5.0000
Calorific value
*L_CAL,*H_CAL 000.000 to 133.000 MJ/m3
0.0000 to 3.5000 kBTU/ft3
Molecular weight
*L_MOL,*H_MOL 000.00 to 140.00
Concentration
*L_CON,*H_CON 000.0 to 100.0
IM 11T03E01-01E
5.1.4
5-5
<5. DENSITY / PARAMETER SETTING>
Setting Parameters at Service Level
Press the [*] key and enter the password (XXX) to select the service level. When you select this
level, the analyzer shows a Code No. promptly. Type the appropriate code number and press the
[ENT] key.
Table 5.4
Code
No.
Setting Parameters at Service Level
Item
01 Setting of hold during maintenance
Hold enable/disable
Setting of preset hold value
02 Setting of hold against errors
Hold enable/disable
Display
*M_HLD
*PR.SET
*E_HLD
Setting of preset hold value
*PR.SET
Data to be Set
(or Conditions)
· Disable : 0
· Enable (value immediately
before) : 1
· Enable (preset value) : 2
-10.0 to 110.0
· Disable : 0
· Enable (value immediately
before) : 1
· Enable (preset value) : 2
-10.0 to 110.0
Remarks
See Figure 5.7 for key operation.
If “preset value” is selected in the
preceding parameter.
See Figure 5.8 for key operation.
If “preset value” is selected in the
preceding parameter.
04 Setting of output-smoothing constants
*SMOTH
00 to 60
See Figure 5.9 for key operation.
05 Setting of contact outputs status
*CNTCT
00 to 15
See Figure 5.10 for key operation.
10 Setting of pressure compensation
Enable/disable/fixed value
*P.COMP
Setting of fixed value for pressure
compensation
11 Setting for compensated density
measurement
Setting of reference temperature
Setting of reference pressure
*P.FIX
See Figure 5.11 for key operation.
· Disable: 0
· Enable (measured value) : 1
· Enable (fixed value) : 2
0.10 to 999.99 kPa, 0.0001 to If “fixed value” is selected in the
9.9999 MPa, 0.015 to 99.999 preceding parameter.
psi
See Figure 5.12 for key operation.
*C.D.TMP
*C.D.PRS
-20.0 to 80.0 °C, - 4.0 to 176.0 °F
0.10 to 999.99 kPa, 0.0001 to
9.9999 MPa, 0.015 to 99.999
psi
12 Setting of gas pressure
Setting of zero gas pressure
*Z_PRS
Setting of span gas pressure
*S_PRS
0.01 to 999.99 kPa, 0.0001 to
9.9999 MPa, 0.015 to 99.999 psi
0.10 to 999.99 kPa, 0.0001 to
9.9999 MPa, 0.015 to 99.999 psi
See Figure 5.13 for key operation.
13 Selection for automatic calibration
*AUTO.C
· Disable : 0
· Enable : 1
See Figure 5.14 for key operation.
14 Setting for remote semi-automatic
calibration
*REMOT
· Disable : 0
· Enable : 1
See Figure 5.15 for key operation.
15 Setting of data for automatic/semiautomatic calibration
Selection of calibration item
Setting of calibration time
Setting of stabilization time
Setting of calibration starting time
Setting of year/month/day
Setting of hour/minute
Selection of unit of calibration
interval
Selection of calibration interval
See Figure 5.16 for key operation.
*CAL.P
*CAL.T
*STAB.T
Zero and span : 0
Zero : 1
Span : 2
00 to 59
00 to 59
*Y_M_D
*H_M
*CYCL.U
00.01.01 to 99.12.31
00.00 to 23.59
Hour : 0
Day : 1
*CYCL.T
Hour : 00 to 23
Day : 000 to 255
IM 11T03E01-01E
Table 5.4
Code
No.
5-6
<5. DENSITY / PARAMETER SETTING>
Setting Parameters at Service Level
Item
Data to be Set
(or Conditions)
Display
Remarks
20 Selection of pressure unit
*PRES.U
· kPa : 0
· MPa : 1
· psi : 2
See Figure 5.17 for key operation.
21 Selection of density unit
*DENS.U
· kg/m3 : 0
· lb/ft3 : 1
See Figure 5.18 for key operation.
· MJ/m3 : 0
· kBTU/ft3 : 1
See Figure 5.19 for key operation.
See Figure 5.20 for key operation.
22 Selection of calorific value unit
*CAL.U
23 Selection of temperature unit
*TEMP.U
· °C : 0
· °F : 1
30 Setting of calendar
Setting of year/month/day
Setting of hour/minute
*Y_M_D
*H_M
00.01.01 to 99.12.31
00.00 to 23.59
*MINUS
· Show : 0
· Hide : 1
31 Showing/hiding of negative measured
values (-)
40 Indication of calibration coefficients
(read-only)
Zero
Span
41 Indication of oscillation frequency
(read-only)
Frequency F2
Frequency F4
Frequency ratio
*C_K_Z
*C_K_S
*F2.KHZ
*F4.KHZ
*F2/F4
*REV
*S_CYC
44 Setting of password
*PASS
45 Selecting Battery alarm detection
50 Selecting the instruments
See Figure 5.23 for key operation.
S of SX.XXXX means Marks such
as a plus or the minus.
*BAT
*MODEL
X.XXXXX
X.XXXXX
X.XXXXX
X.XX
See Figure 5.25 for key operation.
· Normal :
0
· High-resolution : 1
See Figure 5.26 for key operation.
0.0.0 to 9.9.9
See Figure 5.27 for key operation.
· Non detection : 0
· Detection :
1
See Figure 5.28 for key operation.
· Density meter : 0
· Calorie meter : 1
· Hydrogen purity, replacement
meter
:2
See Figure 5.29 for key operation.
82 Setting Detector Constant
5.2
See Figure 5.22 for key operation.
See Figure 5.24 for key operation.
42 Indication of software version
(read-only)
43 Setting of high-resolution mode
SX.XXXX
X.XXXX
See Figure 5.21 for key operation.
For leap years, the setting can
cope with ‘February 29’.
See Figure 5.30 for key operation.
Parameter Setting
[Aborting the Setting Operation]
Press the [MODE] key. Normally, the [MODE] key is used to move from the measurement
() mode to a mode at the operation level. It is also used to return to the
measurement mode from other modes.
5.2.1
Parameter Setting at Operation Level
To move to the operation level, press the [MODE] key and then enter the password (XXX).
Password is not selected in first condition. When password is necessary, refer to Section 4.2
“Setting Lists” and selected password what you need.
IM 11T03E01-01E
5-7
<5. DENSITY / PARAMETER SETTING>
The operation level offers two mode categories: the measurement and calibration operation
modes.
Note that the function for which you have set a data value at the setting level does not work at
all if that function is turned off at another level. For this reason, care must be taken when dealing
with modes that relate to each other.
This subsection explains the operating procedures for the levels/modes noted below.
The calibration operation mode is discussed in Chapter 8.
(1) Operation level
(2) Display mode (DISP mode)
(1) Operation Level
Select a mode from the following four choices of the operation level. Each press of the [NO] key
cycles through the choices in the order shown below:
1. Display mode [DISP]
2. Semi-automatic calibration mode [SEM.CAL]
3. Manual calibration mode [MAN.CAL]
4. Valve operation mode [VALVE]
Press the [YES] key to enter the desired mode when it is indicated.
Measurement Mode
[MODE]
XXX
*PASSW
Password is not selected in first condition
(See "Figure 5.27 Setting of Password" on page 5-31.)
[ENT]
DISP
[NO]
SEM.CAL
[YES]
[YES]
[NO]
MAN.CAL
[YES]
[NO]
VALVE
[YES]
[NO]
DISP mode
See the process in Figure 5.2.
Semi-automatic calibration mode
See the process in Figure 8.3.
Manual calibration mode
See the process in Figure 8.4.
Valve operation mode
See the process in Figure 8.1.
[YES]
[YES]
[YES]
[YES]
F0501.ai
You can abort the setting operation in any of the modes by pressing [MODE] key.
Doing this will return the analyzer to the measurement mode.
Figure 5.1
Modes of Operation Level
IM 11T03E01-01E
<5. DENSITY / PARAMETER SETTING>
5-8
(2) Display Mode
Select one of the following ten parameters to be shown in the message field. Each press of the
[NO] key cycles through the choices in the order shown below:
1. Physical density [XX.XXXX KG/M3] or [X.XXXXX LB/FT3]
(depends on the setting of CODE 21)
2. Compensated density [X.XXXX KG/M3 ]or [X.XXXXX LB/FT3]
(depends on the setting of CODE 21)
3. Specific gravity [X.XXXX SP GR]
4. Calorific value [XXX.XXX MJ/M3] or [X.XXXX KBTU]
(depends on the setting of CODE 22)
5. Molecular weight [XXX.XX MOL]
6. Concentration[XXX.X VOL% ]
7. Temperature [XXX.X °C] or [XXX.X °F]
(depends on the setting of CODE 23)
8. Pressure [XXX.XX KPA] or [X.XXXX MPA] or [XX.XXX PSI]
(depends on the setting of CODE 20)
9. Current for analog output 1 (%) [XXX.X MA1%]
10. Current for analog output 2 (%) [XXX.X MA 2%]
Show the desired choice and press the [YES] key. That choice is set as a parameter to be
measured in the measurement mode (). Parameters 1 to 6 are configurable, while
parameters 7 to 10 are read-only.
IM 11T03E01-01E
*1
<5. DENSITY / PARAMETER SETTING>
Measurement Mode
5-9
[MODE]
DISP
[YES]
XX.XXXX
KG/M3
[YES] *
2,* 3
(Physical density)
[NO]
X.XXXX
[YES] * 2,* 3
KG/M3
(Compensated density)
[NO]
*3
X.XXXX [YES]
SP GR
(Specific gravity)
[NO]
XXX.XXX
MJ/M3
[YES]*
2,* 3
[YES] *
3
(Calorific value)
[NO]
XXX.XX
MOL
(Molecular weight)
[NO]
XXX.X
VOL%
[YES] *
3
[YES] *
2
(Concentration)
[NO]
XXX.X
°C
(Read-only)
(Sample gas temperature)
[NO]
2
[YES] *
XXX.XX
(Read-only)
KPA
(Sample gas pressure)
[NO]
[YES]
XXX.X
(Read-only)
MA1%
(Current for analog output1(%))
[NO]
[YES]
XXX.X
(Read-only)
MA2%
(Current for analog output 2(%))
[NO]
Measurement Mode
F0502.ai
*1: See "Figure 5.1 Modes of Operation Level" on page 5-7.
*2: The unit is the one set in the service level.
*3: Press the [YES] key to show the parameter in the measurement mode.
Figure 5.2
Display Mode (DISP Mode)
IM 11T03E01-01E
5.2.2
<5. DENSITY / PARAMETER SETTING>
5-10
Parameter Setting at Setting Level
To move to the setting level, press the [*] key and then enter the password (XXX). The main task
at the setting level is to set data values such as the measuring range values.
Note that the function for which you have set data values at the setting level, will not work if
that function is turned off at another level. For this reason, care must be taken when dealing
with modes that relate to each other. The setting level has four levels/modes, as shown below,
although this subsection explains only the analog output setting and alarm-point setting modes.
The calibration parameter setting mode is discussed in Chapter 8.
1. Analog output setting mode [*RANGE]
2. Calibration parameter setting mode [*CAL.DT]
3. Alarm-point setting mode [*ALARM]
4. Service level [*SERVC]
Select a mode from the above four choices. Each press of the [NO] key cycles through the
choices in sequence. Press the [YES] key to enter the desired mode when it is indicated.
Measurement Mode
[*]
XXX
*PASSW
Password is not selected in first condition.
(See "Figure 5.27 Process of Setting Password" on page 5-31.)
[ENT]
*RANGE
[NO]
[YES]
See the process in Figure 5.4.
*CAL.DT
[NO]
[YES]
*ALARM
[NO]
Analog output setting mode
[YES]
Calibration parameter setting mode
See the process in Figure 8.2.
Alarm-point setting mode
See the process in Figure 5.5.
*SERVC
[NO]
[YES]
Service level
See the process in Figure 5.7 to 5.30.
[ENT]
[ENT]
[ENT]
[ENT]
F0503.ai
You can abort the setting operation in any of the modes by pressing the [MODE] key.
Doing this will return the analyzer to the measurement mode.
Figure 5.3
Process at the Setting Level
IM 11T03E01-01E
5-11
<5. DENSITY / PARAMETER SETTING>
1. Analog Output Setting Mode [*RANGE]
The analog output has output 1 and output 2. Only output 1 can be used for communication
purposes.
Set a range appropriate for 4-20 mA DC output signals. The range has two set-points: the lower
limit (zero point) of the range corresponding to the minimum (0%) of a given output signal and the
upper limit (span point) of the range corresponding to the maximum (100%) of the given output
signal.
(1) Output1 [*OUT1]
(a) Select a parameter to be output.
Physical density [*DENS]
Compensated density [*C_DNS]
Specific gravity [*SP_GR]
Calorific value [*CALRY]
Molecular weight [*MOL]
Concentration [*CONCT]
Gas temperature [*TEMP]
Gas pressure [*PRESS]
(b) Set the zero and span points. (This item is not applied if temperature or pressure are
chosen as the parameter in item (a) above.)
Physical density [*Z_DNS], [*S_DNS]
Configure range:
00.0000 to 60.0000 (kg/m3),
0.00000 to 4.00000 (lb/ft3)
Compensated density [*Z_CP.D], [*S_CP.D]
Configure range:
0.0000 to 6.0000 (kg/m3),
0.00000 to 0.40000 (lb/ft3)
Specific gravity [*Z_SPC], [*S_SPC]
Configure range:
0.0000 to 5.0000
Calorific value [*Z_CAL], [*S_CAL]
Configure range:
000.000 to 133.000 (MJ/m3),
0.0000 to 3.5000 (kBTU/ft3)
Molecular weight [*Z_MOL], [*S_MOL]
Configure range:
000.00 to 140.00
Concentration [*Z_CON], [*S_CON]
Configure range:
000.0 to 100.0
(2) Output 2 [*OUT2]
The same parameters as those of item (1) apply.
IM 11T03E01-01E
<5. DENSITY / PARAMETER SETTING>
5-12
(3) Entries for Calorie and Concentration Measurement
Measuring the calorific value (or concentration) requires the density to be correlated with the
calorific value (or concentration) by a linear equation (linear proportion).
Therefore, use the gas calorie meter in order to measure the calorific value as accurately as
possible.
The relationship must be given by Y=aX+b, where Y is the calorific value (MJ/m3) and X is the
density (kg/m3).
l Example of Entries for Calorie Measurement
Assuming the relationship is given by “calorific value = 58 × density + 6 (MJ/m3)”, the density
ranges from 0.5 to 0.7 kg/m3 while the calorific value ranges from 35 to 46.6 MJ/m3.
The entries for the [*CALRY] parameter (see Figure 5.4) therefore are:
*Z_CAL: 35
*S_CAL: 46.6
*Z_CL.D: 0.5
*S_CL.D: 0.7
l Example of Entries for Concentration Measurement
Assume we measure the concentration of hydrogen mixed with nitrogen and the relationship
is given by:
D = aX +b
where,
D=density of gas being measured (hydrogen: 0.0899 kg/m3; nitrogen: 1.2504 kg/m3),
and
X=volume percent (vol%) of hydrogen gas,
then the equation above is calculated as:
D= (0.0899-1.2504)(X/100) + 1.2504
= -1.1605(X/100) + 1.2504
1.2504
Density
(D)
X
H2
N2
(1-X)
0.0899
0
100
Volume percent (VOL%)
F0504.ai
The entries for the [*CONCT] parameter (see Figure 5.4) therefore are:
*Z_CON: 0
*S_CON: 100
*Z_CN.D: 1.2504
*S_CN.D: 0.0899
IM 11T03E01-01E
5-13
<5. DENSITY / PARAMETER SETTING>
*1
Measurement mode
XXX
[*]
*PASSW
Password is not selected in first condition.
[ENT]
[NO]
*SERVC
*RANGE
[YES]
*OUT1
[YES]
*2
*DENS
[NO]
[YES]
(Compensated density Range)
[NO]
*SP_GR
(Specific gravity Range)
*3
[NO]
XX.XXXX
[YES]
(Physical density Range)
*C_DNS
[NO]
*Z_DNS
*Z_SPC
[YES]
*S_DNS
[ENT]
X.XXXX
[ENT]
X.XXXX
[YES]
*OUT2
XX.XXXX
[ENT]
X.XXXX
*Z_CP.D
*CAL.DT
*S_CP.D
[ENT]
X.XXXX
[ENT]
*S_SPC
[ENT]
[NO]
*CALRY
XXX.XXX
XXX.XXX
[YES]
*Z_CAL
[YES]
*Z_MOL
[YES]
*Z_CON
(Calorific value Range)
[ENT]
*S_CAL
[ENT]
*S_MOL
[ENT]
*S_CON
X.XXXX
[ENT]
*Z_CL.D
X.XXXX
[ENT]
*S_CL.D
[ENT]
[NO]
*MOL
XXX.XX
XXX.XX
(Molecular weight Range)
[ENT]
[NO]
*CONCT
(Concentration Range)
XXX.X
XXX.X
X.XXXX
[ENT]
*Z_CN.D
X.XXXX
[ENT]
*S_CN.D
[ENT]
[NO]
*TEMP
(Gas temperature)
[YES]
[NO]
*PRESS
(Gas pressure)
[YES]
[NO]
F0505.ai
*1: You can abort the setting operation in any of the modes by pressing [MODE] key.
Doing this will return the analyzer to the measurement mode.
*2: The display shows the previous parameter first.
*3: Each press of the [NO] key toggles between the parameters [*OUT1] and [*OUT2].
Figure 5.4
Process of the Analog Output Setting Mode
IM 11T03E01-01E
5-14
<5. DENSITY / PARAMETER SETTING>
2. Alarm-point Setting Mode [*ALARM]
Set the upper /lower limits for the measured values of the parameters noted below, for the
purpose of raising an alarm.
(a) Select a parameter for which alarms are raised.
Physical density [*DENS]
Compensated density [*C_DNS]
Specific gravity [*SP_GR]
Calorific value [*CALRY]
Molecular weight [*MOL]
Concentration [*CONCT]
Press the [YES] key to enter the desired mode when it is indicated.
(b) Set the upper/lower limits.
Physical density [*L_DNS], [*H_DNS]
Configure range:
00.0000 to 60.0000 (kg/m3),
0.00000 to 4.00000 (lb/ft3)
Compensated density [*L_CP.D], [*H_CP.D]
Configure range:
0.0000 to 6.0000 (kg/m3),
0.00000 to 0.40000 (lb/ft3)
Specific gravity [*L_SPC], [*H_SPC]
Configure range:
0.0000 to 5.0000
Calorific Value [*L_CAL], [*H_CAL]
Configure range:
000.000 to 133.000 (MJ/m3),
0.0000 to 3.5000 (kBTU/ft3)
Molecular weight [*L_MOL], [*H_MOL]
Configure range:
000.00 to 140.00
Concentration [*L_CON], [*H_CON]
Configure range:
000.0 to 100.0
IM 11T03E01-01E
*1
*CAL.DT
<5. DENSITY / PARAMETER SETTING>
[NO]
*ALARM
[NO]
5-15
*SERVC
[YES]
*DENS
(Physical density)
XX.XXXX
[YES]
[NO]
*C_DNS
(Compensated density)
(Specific gravity)
*2
XX.XXXX
[ENT]
X.XXXX
[YES]
[NO]
*SP_GR
*L_DNS
*L_CP.D
*L_SPC
[ENT]
X.XXXX
[ENT]
X.XXXX
[YES]
*H_DNS
*H_CP.D
[ENT]
X.XXXX
[ENT]
*H_SPC
[ENT]
[NO]
*CALRY
(Calorific value)
XXX.XXX
[YES]
*L_CAL
XXX.XXX
[ENT]
*H_CAL
[ENT]
[NO]
*MOL
(Molecular weight)
XXX.XX
[YES]
*L_MOL
XXX.XX
[ENT]
*H_MOL
[ENT]
[NO]
*CONCT
(Concentration)
[YES]
XXX.X
*L_CON
[ENT]
XXX.X
*H_CON
[NO]
[ENT]
F0506.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: The display shows the previous parameter first.
Figure 5.5
Process of the Alarm-point Setting Mode
IM 11T03E01-01E
5.2.3
<5. DENSITY / PARAMETER SETTING>
5-16
Parameter Setting at Service Level
To move to the service level, select [*SERVC] at the setting level, and press the [YES] key.
(see Figure 5.3 on page 5-10.) The main task at the setting level is to set data values such as a
measuring range. To go into each mode of the service level, enter the appropriate code number.
CAUTION
DO NOT enter codes other than those listed in Table 5.4. A change to the settings of any unlisted
code can result in the failure of the analyzer to operate properly. If you have entered a wrong code
number by mistake, press the [MODE] key to return to the measurement mode().
You may exit the mode of that wrong code by pressing the [ENT] key, if you haven't made any
change to the relevant data.
(1)
When [*SERVC] is shown on the display, press
[YES] key. The message field changes to look
like the one shown left.
(2)
Press the [ENT] key with code number 01 shown
on the display, for example. The analyzer enters
the mode for setting the parameters of that code.
F0507.ai
Figure 5.6
Entry of Code Number at Service Level
The following explains the parameter setting procedures in the order of code numbers.
IM 11T03E01-01E
<5. DENSITY / PARAMETER SETTING>
5-17
CODE 01: Setting of Hold during Maintenance
This setting determines whether or not the output signal is held during maintenance. As a value
to be held, you can select either a value immediately before or a preset value.
(a) Hold enable/disable [*M_HLD]
Disable:
0
Enable:
1
Enable (preset value): 2
(b) Setting of preset hold value [*PR.SET]
This item applies if item (a) above is set to "2".
Configurable range:
-10.0 to 110.0 (percent of analog output)
*1
*SERVC
[YES]
01
*CODE
*2
[ENT]
X
*M_HLD
*5
*3
[ENT]
XXX.X
*PR.SET
*4
[ENT]
F0508.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value using the [ ] and [ENT] keys.
*4: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*5: Setting 0 or 1 returns to the service level with the indication [*SERVC].
Figure 5.7
Setting of Hold during Maintenance
IM 11T03E01-01E
<5. DENSITY / PARAMETER SETTING>
5-18
CODE 02: Setting of Hold in the Event of Error
This setting determines whether or not the output signal is held if an error occurs. You can select
either the value immediately prior to the error or a preset value as the value to be held.
(a) Hold enable/disable [*E_HLD]
Disable:
0
Enable:
1
Enable (preset value):
2
(b) Setting of preset hold value [*PR.SET]
This item applies if item (a) above is set to "2".
Configurable range: -10.0 to 110.0 (percent of analog output)
*1
*SERVC
[YES]
02
*CODE
*2
[ENT]
X
*E_HLD
*5
*3
[ENT]
XXX.X
*PR.SET
*4
[ENT]
F0509.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value using the [ ] and [ENT] keys.
*4: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*5: Setting 0 or 1 returns to the service level with the indication [*SERVC].
Figure 5.8
Setting of Hold in the Event of Error
CODE 04: Setting of Output-smoothing Constants [*SMOTH]
This setting defines the constants for output smoothing. This smoothing constants is outputsmoothing constants of converter (electrical circuit boards).
Configurable range: 00 to 60 (unit : second)
*1
*SERVC
[YES]
04
*CODE
*2
[ENT]
XX
*SMOTH
*3
[ENT]
F0510.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
Figure 5.9
Setting of Output-smoothing Constants
IM 11T03E01-01E
<5. DENSITY / PARAMETER SETTING>
5-19
CODE 05: Setting of Contact Outputs Status [*CNTCT]
This setting defines the status of the contact outputs.
Configurable range: 00 to 15
*1
*SERVC
[YES]
05
*CODE
*2
[ENT]
XX
*CNTCT
*3
[ENT]
F0511.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
Figure 5.10
Setting of Contact Outputs Status
Each contact output takes either of the following two status (in table), depending on the value you
set.
Value
ZERO contact
SPAN contact
MAINT contact
ALM contact
0
NO
NO
NO
NO
1
NO
NO
NO
NC
2
NO
NO
NC
NO
3
NO
NO
NC
NC
4
NO
NC
NO
NO
5
NO
NC
NO
NC
6
NO
NC
NC
NO
7
NO
NC
NC
NC
8
NC
NO
NO
NO
9
NC
NO
NO
NC
10
NC
NO
NC
NO
11
NC
NO
NC
NC
12
NC
NC
NO
NO
13
NC
NC
NO
NC
14
NC
NC
NC
NO
15
NC
NC
NC
NC
NO: means “normally open”
NC: means “normally closed”
IM 11T03E01-01E
<5. DENSITY / PARAMETER SETTING>
5-20
CODE 10: Setting of Pressure Compensation
This setting determines whether or not the measured density should be compensated by the gas
pressure. The density can be compensated to either the value immediately before the error or a
fixed value.
(a) Enable/disable/fixed value [*P.COMP]
Disable: 0
Enable (value immediately before): 1
Enable (fixed value):
2
(b) Setting of fixed value for pressure compensation [*P.FIX]
This item applies if item (a) above is set to "2".
Configurable range: 0.10 to 999.99 (for kPa),
0.0001 to 9.9999 (for MPa),
0.015 to 99.999 (for psi )
*1
*SERVC
[YES]
10
*CODE
*2
[ENT]
X
*P.COMP
*5
*3
[ENT]
XXX.XX
*P.FIX
*6
*4
[ENT]
F0512.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value using the [ ] and [ENT] keys.
*4: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*5: Setting 0 or 1 returns to the service level with the indication [*SERVC].
*6: The position of the decimal point changes depending on the setting of CODE 20.
If the code is set to "0" (kPa), the reading is XXX.XX;
if the code is set to "1" (MPa), the reading is X.XXXX;
if the code is set to "2" (psi), the reading is XX.XXX.
Figure 5.11
Setting of Pressure Compensation
IM 11T03E01-01E
<5. DENSITY / PARAMETER SETTING>
5-21
CODE 11: Setting for Compensated Density Measurement
This setting defines the reference temperature and pressure required to obtain the compensated
density.
(a) Setting of reference temperature [*C.D.TMP]
Configurable range:
-20.0 to 80.0 (for °C)
-4.0 to 176.0 (for °F )
(b) Setting of reference pressure [*C.D.PRS]
Configurable range:
*1
*SERVC
[YES]
0.10 to 999.99 (for kPa),
0.0001 to 9.9999 (for MPa),
0.015 to 99.999 (for psi )
11
*CODE
*2
[ENT]
XX.X
*C.D.TMP
*4
*3
[ENT]
XXX.XX
*C.D.PRS
*5
*3
[ENT]
F0513.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*4: The position of the decimal point changes depending on the setting of CODE 23.
If the code is set to "0" (°C), the reading is XX.X;
if the code is set to "1" (°F), the reading is XXX.X.
*5: The position of the decimal point changes depending on the setting of CODE 20.
If the code is set to "0" (kPa), the reading is XXX.XX;
if the code is set to "1" (MPa), the reading is X.XXXX;
if the code is set to "2" (psi), the reading is XX.XXX.
Figure 5.12
Setting for Compensated Density Measurement
IM 11T03E01-01E
<5. DENSITY / PARAMETER SETTING>
5-22
CODE 12: Setting of Gas Pressure
This setting defines the zero and span points of the pressure transmitter's measurement range.
(a) Setting of zero-point gas pressure [*Z_PRS]
Configurable range:
0.01 to 999.99 (for kPa),
0.0001 to 9.9999 (for MPa),
0.015 to 99.999 (for psi )
(b) Setting of span-point gas pressure [*S_PRS]
Configurable range: 0.10 to 999.99 (for kPa),
0.0001 to 9.9999 (for MPa),
0.015 to 99.999 (for psi )
The zero and span points set here are also used for the analog output.
*1
*SERVC
[YES]
12
*CODE
*2
[ENT]
XXX.XX
*Z_PRS
*4
*3
[ENT]
XXX.XX
*S_PRS
*4
*3
[ENT]
F0514.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*4: The position of the decimal point changes depending on the setting of CODE 20.
If the code is set to "0" (kPa), the reading is XXX.XX;
if the code is set to "1" (MPa), the reading is X.XXXX;
if the code is set to "2" (psi), the reading is XX.XXX.
Figure 5.13
Setting of Gas Pressure
IM 11T03E01-01E
5-23
<5. DENSITY / PARAMETER SETTING>
CODE 13: Selection for Automatic Calibration [*AUTO.C]
This setting determines whether or not automatic calibration is carried out. If automatic calibration
is to be carried out, you must also configure CODE 15 at the service level.
Disable:
0
Enable:
1
*1
*SERVC
[YES]
13
*CODE
*2
[ENT]
X
*AUTO.C
*3
[ENT]
F0515.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
Figure 5.14
Selection for Automatic Calibration
CODE 14: Setting for Remote Semi-automatic Calibration [*REMOT]
This setting determines whether or not semi-automatic calibration is carried out remotely.
If remote semi-automatic calibration is to be carried out, you must wire the contact input.
You must also configure CODE 15 at the service level.
Disable:
0 (The contact is open)
Enable:
1 (The contact is closed)
*1
*SERVC
[YES]
14
*CODE
*2
[ENT]
X
*REMOT
*3
[ENT]
F0516.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
Figure 5.15
Setting for Remote Semi-automatic Calibration
IM 11T03E01-01E
<5. DENSITY / PARAMETER SETTING>
5-24
CODE 15: Setting of Data for Automatic/Semi-automatic Calibration
This setting defines the data needed to carry out automatic/semi-automatic calibration.
(a) Selection of calibration item [*CAL.P]
Zero and span:
0
Zero:
1
Span:
2
(b) Setting of calibration time [*CAL.T]
Configurable range: 00 to 59 (unit : minute)
(c) Setting of stabilization time [*STAB.T]
Configurable range: 00 to 59 (unit : minute)
(d) Setting of calibration starting time
(1) Setting of year/month/day [*Y_M_D]
Configurable range: 00.01.01 to 99.12.31
(2) Setting of hour/minute [*H_M]
Configurable range: 00.00 to 23.59
(e) Selection of unit of calibration interval [*CYCL.U]
Hour: 0
Day: 1
(f) Selection of calibration interval [*CYCL.T]
Item (e) above is set to "0".
Configurable range: 00 to 23 (for hour)
Item (e) above is set to "1".
Configurable range: 000 to 255 (for day)
For more information, see Chapter 8, “Calibration Procedure”.
IM 11T03E01-01E
5-25
<5. DENSITY / PARAMETER SETTING>
*1
*SERVC
[YES]
15
*CODE
*2
[ENT]
X
*CAL.P
*2
(Selection of calibration item)
[ENT]
XX
*CAL.T
*2
(Setting of calibration time)
[ENT]
XX
*STAB.T
*2
(Setting of stabilization time)
[ENT]
XX.XX.XX
*Y_M_D
*2
(Setting of calibration starting time [year/ month/ day])
[ENT]
XX.XX
*H_M
*2
(Setting of calibration starting time [hour/ minute])
[ENT]
X
*CYCL.U
*2
(Selection of unit of calibration interval)
[ENT]
XXX
*CYCL.T
*3
*2
(Selection of calibration interval)
[ENT]
F0517.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*3: The number of digits on LCD varies depending on the unit of the calibration interval (parameter [*CYCL.U]).
If the parameter is set to "0" (hour), the indication is two-digit number.
If the parameter is set to "1" (day), the indication is three-digit number.
Figure 5.16
Setting of Data for Automatic/Semi-automatic Calibration
IM 11T03E01-01E
<5. DENSITY / PARAMETER SETTING>
5-26
CODE 20: Selection of Pressure Unit [*PRES.U]
This setting defines the pressure unit for the gas pressure transmitter.
kPa:
0
MPa: 1
psi :
2
*1
*SERVC
[YES]
20
*CODE
*2
[ENT]
X
*PRES.U
*3
[ENT]
F0518.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
Figure 5.17
Selection of Pressure Unit.
CODE 21: Selection of Density Unit [*DENS.U]
This setting defines the density unit for the gas density meter.
kg/m3: 0
lb/ft3:
1
*1
*SERVC
[YES]
21
*CODE
*2
[ENT]
X
*DENS.U
*3
[ENT]
F0519.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
Figure 5.18
Selection of Density Unit
IM 11T03E01-01E
<5. DENSITY / PARAMETER SETTING>
5-27
CODE 22: Selection of Calorific Value Unit [ *CAL.U]
This setting defines the calorific value unit for the gas density meter.
• MJ/m3 :
0
• kBTU/ft3 :
*1
*SERVC
[YES]
1
22
*CODE
*2
[ENT]
X
*CAL.U
*3
[ENT]
F0520.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
Figure 5.19
Selection of Calorific Value Unit
CODE 23: Selection of Temperature Unit [*TEMP.U]
This setting defines the temperature unit for the gas density meter.
• °C:
0
• °F:
1
*1
*SERVC
[YES]
23
*CODE
*2
[ENT]
X
*TEMP.U
*3
[ENT]
F0521.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
Figure 5.20
Selection of Temperature Unit
IM 11T03E01-01E
5-28
<5. DENSITY / PARAMETER SETTING>
CODE 30: Setting of Calendar
This setting defines the date and time.
(a) Setting of year/month/day [*Y_M_D]
Configurable range : 00.01.01 to 99.12.31
(b) Setting of hour/minute [*H_M]
Configurable range : 00.00 to 23.59
*1
*SERVC
[YES]
30
*CODE
*2
[ENT]
XX.XX.XX
*Y_M_D
*3
[ENT]
XX.XX
*H_M
*3
[ENT]
F0522.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
Figure 5.21
Setting of Calendar
CODE 31: Showing/Hiding of Negative Measured Values [*MINUS]
This setting determines whether negative measured values (-) should be shown or hidden.
Show: 0
Hide: 1
*1
*SERVC
[YES]
31
*CODE
*2
[ENT]
X
*MINUS
*3
[ENT]
F0523.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
Figure 5.22
Showing/Hiding of Negative Measured Values
IM 11T03E01-01E
5-29
<5. DENSITY / PARAMETER SETTING>
CODE 40: Indication of Calibration Coefficients (Read-only)
This setting determines the calibration coefficients to be indicated.
(a) Zero [*C_K_Z]:
indicated as SX.XXXX
(b) Span [*C_K_S]:
indicated as X.XXXX
*1
*SERVC
[YES]
40
*CODE
*2
[ENT]
SX.XXXX
*C_K_Z
*3
[ENT]
X.XXXX
*C_K_S
[ENT]
F0524.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: By pressing the [ENT] key, the next indication is displayed.
This parameter does not accept the [ ] and [ ] key input.
Figure 5.23
Indication of Calibration Coefficients
CODE 41: Indication of Oscillation Frequency (Read-only)
This setting determines the F2 and F4 oscillation frequencies coefficients to be indicated.
(a) Indication of frequency F2 (kHz) [*F2.KHZ]:
indicated as X.XXXXX
(b) Indication of frequency F4 (kHz) [*F4.KHZ]:
indicated as X.XXXXX
(c) Indication of frequency ratio (kHz) [*F2/F4]:
indicated as X.XXXXX
*1
*SERVC
[YES]
*2
41
*CODE
[ENT]
X.XXXXX
*F2.KHZ
*3
[ENT]
X.XXXXX
*F4.KHZ
*3
[ENT]
X.XXXXX
*F2/F4
*3
[ENT]
F0525.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: By pressing the [ENT] key, the next indication is displayed.
This parameter does not accept the [ ] and [ ] key input.
Figure 5.24
Indication of Oscillation Frequency
IM 11T03E01-01E
<5. DENSITY / PARAMETER SETTING>
5-30
CODE 42: Indication of Software Version [*REV] (Read-only)
This setting allows the software version to be confirmed (indicated).
Indicated as X.XX
*1
*SERVC
[YES]
42
*CODE
*2
[ENT]
X.XX
*REV
*3
[ENT]
F0526.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Pressing the [ENT] key returns to the service level with the indication [*SERVC].
This parameter does not accept the [ ] and [ ] key input.
Figure 5.25
Indication of Software Version
CODE 43: Setting of High-resolution Mode [*S_CYC]
This setting selects between the normal mode and the high-resolution mode.
Normal mode:
0
High-resolution mode: 1
*1
*SERVC
[YES]
43
*CODE
*2
[ENT]
X
*S_CYC
*3
[ENT]
F0527.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value by using the [ ] and [ ] key.
Figure 5.26
Setting of High-resolution Mode
IM 11T03E01-01E
<5. DENSITY / PARAMETER SETTING>
5-31
CODE 44: Setting of PASSWORD [ *PASS]
This setting defines the password.
*1
*SERVC
[YES]
44
*CODE
*2
[ENT]
X.X.X
*PASS
*3
[ENT]
F0528.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value by using the [ ], [ ] and [ENT] keys.
X.X.X
Service level
Setting level
Operation level
Note: Password entry request ( if a password is set)
At the operation level, a request is issued when the
[MODE] key is pressed in the measurement mode. At the
setting level, it is issued when the [ * ] key is pressed. At
the service level, it is issued when the [YES] key is
pressed with [*SERVC] displayed.
Figure 5.27
Password
X
(Setting)
Contents
(0)
1
2
3
4
5
6
7
8
9
(No setting)
111
333
777
888
123
957
331
546
847
Setting of Password
CODE 45: Selecting Battery alarm detection [ *BAT]
This setting determines whether or not battery alarm is detection.
Non detection: 0
Detection:
1
*1
*SERVC
[YES]
45
*CODE
*2
[ENT]
X
*BAT
*3
[ENT]
F0529.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value by using the [ ], [ ] and [ENT] keys.
Figure 5.28
Selecting Battery alarm detection
IM 11T03E01-01E
<5. DENSITY / PARAMETER SETTING>
5-32
CODE 50: Selecting the instruments [ *MODEL]
This setting defines the model.
Density meter:
Calorie meter:
Hydrogen purity, replacement meter :
*1
*SERVC
[YES]
50
*CODE
0
1
2
*2
[ENT]
X
*MODEL
*3
[ENT]
F0530.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value by using the [ ], [ ] and [ENT] keys.
Figure 5.29
Selecting the instruments
CODE 82: Setting Detector Constant
This setting defines the detector constant.
CAUTION
The instruments may not only work improperly but also big problems occur unless the detector
constants entered incorrectly. Do carefully and precisely when enter the detector constants into
converter.
1. The GD402 gas density meter is shipped after adjusting the detector and converter in pairs.
When installation, confirm the converter serial number described on the label of detector so
that combines converter and detector correctly. If mismatched in pairs, converter may be out
of order. If combined correctly, no need to enter the detector constants again, as converter
has been adjusted with the constants in factory.
2. When supply converter or detector individually, enter the detector constants, described on
inside the lid of the GD40, into the converter so that the GD402 is going to be well.
IM 11T03E01-01E
*1
<5. DENSITY / PARAMETER SETTING>
[YES]
*SERVC
82
*CODE
5-33
*2
[ENT]
020
*PASSW
*4
[ENT]
XXXX
*K_A_H
*3
[ENT]
XXXX
*K_A_L
*3
[ENT]
XXXX
*K_B_H
*3
[ENT]
XXXX
*T2L_H
*3
[ENT]
XXXX
*T2L_L
*3
[ENT]
XXXX
*T3L_H
*3
[ENT]
XXXX
*T3L_L
*3
[ENT]
XXXX
*3
*K_B_L
[ENT]
XXXX
*3
*T1H_H
[ENT]
XXXX
*K_C_H
XXXX
*T1H_L
*3
[ENT]
XXXX
*K_C_L
*3
[ENT]
XXXX
*K_AH.H
*3
[ENT]
XXXX
*K_AH.L
*3
[ENT]
XXXX
*K_BH.H
*3
[ENT]
XXXX
*K_BH.L
*3
[ENT]
XXXX
*K_CH.H
*3
[ENT]
XXXX
*K_CH.L
*3
[ENT]
XXXX
*T1L_H
*3
[ENT]
XXXX
*T1L_L
*3
[ENT]
*3
[ENT]
XXXX
*T2H_H
*3
[ENT]
XXXX
*T2H_L
*3
[ENT]
XXXX
*T3H_H
*3
[ENT]
XXXX
*T3H_L
*3
[ENT]
XX.XX
*M_T1
*3
[ENT]
XX.XX
*M_T2
*3
[ENT]
XX.XX
*C_T
*3
[ENT]
XX.XXX
*TMP_Z
*3
[ENT]
XX.XXX
*TMP_S
*3
[ENT]
F0531.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
Detector Constants are described inside the lid of detector GD40.
*4: Set the PASSWORD.
Figure 5.30 Setting Detector Constant
IM 11T03E01-01E
Blank Page
6.
6-1
<6. CALORIE / PARAMETER SETTING>
CALORIE / PARAMETER SETTING
CAUTION
Password is not selected in first condition. When password is necessary, refer to Section “4.2
Setting Lists”.
This chapter describes how to set parameters of Calorie meter.
If select Density meter, see Chapter 5.
If select Hydrogen purity meter, see Chapter 7.
Conversion Table
1 lb/ft3
1 kBTU/ft3
= 16.01847 kg/m3
= 37.25901232 MJ/m3
X degF
= (X-32)/1.8
1 psi
= 6.894756794 kPa
Density of Air = 1.2928 kg/m3
= 0.08070684 lb/ft3
IM 11T03E01-01E
6.1
6-2
<6. CALORIE / PARAMETER SETTING>
Setting Parameters
Subsections 6.1.1 to 6.1.4 show the setting parameters for each level.
6.1.1
Setting Parameters at Measurement Level
When turned on, the analyzer starts up in the measurement mode ().
Table 6.1
Setting Parameters at Measurement Level
Mode/Setting parameter
Measurement mode
Indication of measurement value
Physical density
Compensated density
Calorific value
Indication of message
Unit
Error symbol and number
Alarm mark and number
Indication of HOLD and FAIL
HOLD
FAIL
DISP mode
Physical density
Compensated density
Calorific value
Sample gas temperature
Sample gas pressure
Analog output 1 (%)
Analog output 2 (%)
Display
Data to be Set
(or Conditions)
Remarks
See Figure 6.3 for key operation.
KG/M3
LB/FT3
KG/M3
LB/FT3
MJ/M3
KBTU
Set the unit in CODE 21.
Set the unit in CODE 21.
Set the unit in CODE 22.
The unit is the one set in the service level.
Appears if an error occurs.
Appears if an alarm occurs.
Appears even during other modes.
Appears during a HOLD state.
Appears if a failure occurs.
ERR.NO
ALM.NO
HOLD
FAIL
DISP
KG/M3
LB/FT3
KG/M3
LB/FT3
MJ/M3
KBTU
ºC
ºF
KPA
MPA
PSI
MA1%
MA2%
Set the unit in CODE 21.
Set the unit in CODE 21.
See Figure 6.3 for key operation.
Selection of parameters to be measured in
the measurement mode.
Set the unit in CODE 22.
Set the unit in CODE 23.
Read-only
Set the unit in CODE 20.
Read-only
Read-only
Read-only
IM 11T03E01-01E
6.1.2
6-3
<6. CALORIE / PARAMETER SETTING>
Setting Parameters at Operation Level
Press the [MODE] key and enter the password (XXX) to gain access to this level (see Figure 6.1
for key operation).
Table 6.2
Setting Parameters at Operation Level
Mode/Setting parameter
Calibration mode
Semi-automatic calibration
Manual calibration
Setting of zero-point calorific
value
Setting of zero-point density
Setting of span-point calorific
value
Setting of span-point density
Zero/span-gas valve operation
Display
SEM.CAL
MAN.CAL
Z_CAL
See Figure 9.3 for key operation.
See Figure 9.4 for key operation.
S_DNS
VALVE
Sample-gas valve operation
Automatic zero-point adjustment
Factor setting
Zero point
Factor K0
Factor K1
One-touch calibration
Remarks
Z_DNS
S_CAL
Zero-gas valve operation
Span-gas valve operation
Density/calorie conversion factor
setting mode
Data to be Set
(or Conditions)
ADJUST
AT_ADJ
SET_K
K0_ADJ
K0
K1
No setting is required for See Figure 9.1 for key operation.
manual
calibration.
Control of zero-gas valve
Control of span-gas
valve
Switching between
sample and
calibration gases
The decimal point can
be repositioned for this
setting only.
Range : -99999 to 99999
Range : -99999 to 99999
Range : 0 to 99999
See Figure 6.2 for key operation.
(Floating-point)
IM 11T03E01-01E
6.1.3
6-4
<6. CALORIE / PARAMETER SETTING>
Setting Parameters at Setting Level
Press the [*] key and enter the password (XXX) to gain access to this level (see Figure 6.5 for key
operation).
Table 6.3
Setting Parameters at Setting Level
Mode/Setting parameter
Analog output setting
Output 1
Set a parameter to be output
Set the zero and span points
Physical density
Compensated density
Calorific value
Output 2 (same as output 1)
Display
*RANGE
*OUT1
*DENS
*C_DNS
*CALRY
*TEMP
*PRESS
*Z_DNS,
*S_DNS
*Z_CP.D,
*S_CP.D
*Z_CAL,
*S_CAL
*OUT2
Setting calibration data
Zero-point calorific value
Zero-point density
*CAL.DT
*Z_CAL
*Z_DNS
Span-point calorific value
*S_CAL
Span-point density
*S_DNS
Output hold value during
calibration
*C_HLD
Setting of preset value
(Only if the preceding parameter
is set to 2 (preset value))
*PR.SET
Alarm-point setting
Select a parameter for which an
alarm is set
Physical density
Compensated density
Calorific value
Set the high/low limits.
Physical density
Compensated density
Calorific value
Data to be Set
(or Conditions)
Remarks
See Figure 6.5 for key operation.
Physical density
Compensated density
Calorific value
Gas temperature
Gas pressure
00.0000 to 60.0000 kg/m3 No setting is required if the parameter is
0.00000 to 4.00000 lb/ft3 temperature or pressure.
0.0000 to 6.0000 kg/m3
0.00000 to 0.40000 lb/ft3
000.000 to 133.000 MJ/m3
0.0000 to 3.5000 kBTU/ft3
000.000 to 133.000 MJ/m3
0.0000 to 6.0000 kg/m3
0.00000 to 0.40000 lb/ft3
000.000 to 133.000 MJ/m3
0.0000 to 3.5000 kBTU/ft3
0.0000 to 6.0000 kg/m3
0.00000 to 0.40000 lb/ft3
· Disable: 0
· Enable (value
immediately before) : 1
· Enable (preset value) : 2
-10.0 to 110.0
*ALARM
See Figure 9.2 for key operation.
See Figure 6.6 for key operation.
*DENS
*C_DNS
*CALRY
*L_DNS,
*H_DNS
*L_CP.D,
*H_CP.D
*L_CAL,
*H_CAL
00.0000 to 60.0000 kg/m3
0.0000 to 6.0000 kg/m3
000.000 to 133.000 MJ/m3
IM 11T03E01-01E
6.1.4
6-5
<6. CALORIE / PARAMETER SETTING>
Setting Parameters at Service Level
Press the [*] key and enter the password (XXX) to select the service level. When you select this
level, the analyzer shows a Code No. promptly. Type the appropriate code number and press the
[ENT] key.
Table 6.4
Code
No.
Setting Parameters at Service Level
Item
01 Setting of hold during maintenance
Hold enable/disable
Setting of preset hold value
02 Setting of hold against errors
Hold enable/disable
Display
*M_HLD
*PR.SET
· Disable : 0
· Enable (value immediately
before) : 1
· Enable (preset value) : 2
-10.0 to 110.0
Setting of preset hold value
*PR.SET
· Disable : 0
· Enable (value immediately
before) : 1
· Enable (preset value) : 2
-10.0 to 110.0
04 Setting of output-smoothing constants
*SMOTH
00 to 60
05 Setting of contact outputs status
*CNTCT
00 to 15
10 Setting of pressure compensation
Enable/disable/fixed value
Setting of fixed value for pressure
compensation
11 Setting for compensated density
measurement
Setting of reference temperature
Setting of reference pressure
*E_HLD
Data to be Set
(or Conditions)
*P.COMP
*P.FIX
Remarks
See Figure 6.8 for key operation.
If “preset value” is selected in the
preceding parameter.
See Figure 6.9 for key operation.
If “preset value” is selected in the
preceding parameter.
See Figure 6.10 for key operation.
See Figure 6.11 for key operation.
See Figure 6.12 for key operation.
· Disable: 0
· Enable (measured value) : 1
· Enable (fixed value) : 2
0.10 to 999.99 kPa, 0.0001 to
If “fixed value” is selected in the
9.9999 MPa, 0.015 to 99.999 psi preceding parameter.
See Figure 6.13 for key operation.
*C.D.TMP
*C.D.PRS
12 Setting of gas pressure
Setting of zero-gas pressure
*Z_PRS
Setting of span-gas pressure
*S_PRS
-20.0 to 80.0°C, -4.0 to 176.0°F
0.10 to 999.99 kPa, 0.0001 to
9.9999 MPa, 0.015 to 99.999 psi
0.01 to 999.99kPa, 0.0001 to
9.9999MPa, 0.015 to 99.999 psi
0.10 to 999.99kPa, 0.0001 to
9.9999MPa, 0.015 to 99.999 psi
See Figure 6.14 for key operation.
13 Selection for automatic calibration
*AUTO.C
· Disable : 0
· Enable : 1
See Figure 6.15 for key operation.
14 Setting for remote semi-automatic
calibration
*REMOT
· Disable : 0
· Enable : 1
See Figure 6.16 for key operation.
15 Setting of data for automatic/semiautomatic calibration
Selection of calibration item
Setting of calibration time
Setting of stabilization time
Setting of calibration starting time
Setting of year/month/day
Setting of hour/minute
Selection of unit of calibration
interval
Selection of calibration interval
See Figure 6.17 for key operation.
*CAL.P
*CAL.T
*STAB.T
*Y_M_D
*H_M
*CYCL.U
*CYCL.T
Zero and span : 0
Zero : 1
Span : 2
00 to 59
00 to 59
00.01.01 to 99.12.31
00.00 to 23.59
Hour : 0
Day : 1
Hour : 00 to 23
Day : 000 to 255
IM 11T03E01-01E
Table 6.4
Code
No.
6-6
<6. CALORIE / PARAMETER SETTING>
Setting Parameters at Service Level
Item
Data to be Set
(or Conditions)
Display
Remarks
20 Selection of pressure unit
*PRES.U
· kPa : 0
· MPa : 1
· psi : 2
See Figure 6.18 for key operation.
21 Selection of density unit
*DENS.U
· kg/m3 : 0
· lb/ft3 : 1
See Figure 6.19 for key operation.
· MJ/m3 : 0
· kBTU/ft3 : 1
See Figure 6.20 for key operation.
See Figure 6.21 for key operation.
22 Selection of calorific value unit
*CAL.U
23 Selection of temperature unit
*TEMP.U
· °C : 0
· °F : 1
30 Setting of calendar
Setting of year/month/day
Setting of hour/minute
*Y_M_D
*H_M
00.01.01 to 99.12.31
00.00 to 23.59
*MINUS
· Show : 0
· Hide : 1
31 Showing/hiding of negative measured
values (-)
40 Indication of calibration coefficients
(read-only)
Zero
Span
41 Indication of oscillation frequency
(read-only)
Frequency F2
Frequency F4
Frequency ratio
42 Indication of software version
(read-only)
43 Setting of high-resolution mode
44 Setting of password
45 Selecting Battery alarm detection
50 Selecting the instruments
82 Setting Detector Constant
*C_K_Z
*C_K_S
SX.XXXX
X.XXXX
See Figure 6.22 for key operation.
For leap years, the setting can
cope with ‘February 29’.
See Figure 6.23 for key operation.
See Figure 6.24 for key operation.
S of SX.XXXX means Marks such
as a plus or the minus.
See Figure 6.25 for key operation.
*F2.KHZ
*F4.KHZ
*F2/F4
*REV
*S_CYC
*PASS
*BAT
*MODEL
X.XXXXX
X.XXXXX
X.XXXXX
X.XX
See Figure 6.26 for key operation.
· Normal :
0
· High-resolution : 1
See Figure 6.27 for key operation.
0.0.0 to 9.9.9
See Figure 6.28 for key operation.
· Non detection : 0
· Detection :
1
See Figure 6.29 for key operation.
· Density meter : 0
· Calorie meter : 1
· Hydrogen purity, replacement
meter :
2
See Figure 6.30 for key operation.
See Figure 6.31 for key operation.
IM 11T03E01-01E
6.2
6-7
<6. CALORIE / PARAMETER SETTING>
Parameter Setting
[Aborting the Setting Operation]
Press the [MODE] key. Normally, the [MODE] key is used to move from the measurement
() mode to a mode at the operation level. It is also used to return to the
measurement mode from other modes.
6.2.1
Parameter Setting at Operation Level
To move to the operation level, press the [MODE] key and then enter the password (XXX).
Password is not selected in first condition. When password is necessary, refer to Section 4.2
“Setting Lists” and selected password what you need.
The operation level offers two mode categories: the measurement and calibration operation
modes.
Note that the function for which you have set a data value at the setting level does not work at
all if that function is turned off at another level. For this reason, care must be taken when dealing
with modes that relate to each other.
This subsection explains the operating procedures for the levels/modes noted below.
The calibration operation mode is discussed in Chapter 9.
(1) Operation level
(2) Density/calorie conversion factor setting mode
(3) Display mode (DISP mode)
IM 11T03E01-01E
6-8
<6. CALORIE / PARAMETER SETTING>
(1) Operation Level
Select a mode from the following four choices of the operation level. Each press of the [NO] key
cycles through the choices in the order shown below:
1. Display mode [DISP]
2. Semi-automatic calibration mode [SEM.CAL]
3. Manual calibration mode [MAN.CAL]
In the mode, “Density / Calorie Conversion Factor Setting Mode” is included in next page.
4. Valve operation mode [VALVE]
Press the [YES] key to enter the desired mode when it is indicated.
Measurement Mode
[MODE]
XXX
*PASSW
Password is not selected in first condition
(See "Figure 6.28 Setting of Password" on page 6-31.)
[ENT]
DISP
[NO]
SEM.CAL
[YES]
[YES]
[NO]
MAN.CAL
[YES]
[NO]
VALVE
[YES]
[NO]
DISP mode
See the process in Figure 6.3.
Semi-automatic calibration mode
See the process in Figure 9.3.
Manual calibration mode
See the process in Figure 9.4.
Valve operation mode
See the process in Figure 9.1.
[YES]
[YES]
[YES]
[YES]
F0601.ai
You can abort the setting operation in any of the modes by pressing [MODE] key.
Doing this will return the analyzer to the measurement mode.
Figure 6.1
Modes of Operation Level
IM 11T03E01-01E
<6. CALORIE / PARAMETER SETTING>
6-9
(2) Density/Calorie Conversion Factor Setting Mode
In order to this mode, select “MAN.CAL” in before page, and press [YES] key.
This mode is used to set a factor for converting density to calorie. This setting is mandatory when
measuring the calorie.
(a) Automatic zero-point adjustment [AT_ADJ] : 0 to 133.000, 0 to 3.5000
(b) Factor setting [SET_K] :
• Zero point [K0_ADJ] : -99999 to 99999
• K0 [K0] :
-99999 to 99999
• K1 [K1] :
0 to 99999
Note: The result of executing item (a) above is also reflected on [K0_ADJ] in item (b).
*1
MAN.CAL
[YES]
ZERO
[NO]
SPAN
[NO]
Measurement mode
ADJUST
[YES]
[YES]
[NO]
AT_ADJ
SET_K
[YES]
XXX.XXX
VALVE
[YES]
END
[NO]
X.XXXX
*3
K0_ADJ
[ENT]
XXX.XXX
CAL.SET
[NO]
*2
[ENT]
X.XXXX
*4
K0
*2
[ENT]
[ENT]
X.XXXX
K1
*2
[ENT]
*1: See "Figure 6.1 Modes of Operation Level" on page 6-8.
*2: Using the [ ], [ ] and [ENT] keys, set a value (numeral position of the decimal point and sign).
Each press of the [ ] key moves through the fields of the numeral decimal point and sign in this order.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*3: Set a value using the [ ], [ ] and [ENT] keys.
*4 When the reading stabilizes, press the [ENT] key to confirm the factor.
F0602.ai
This parameter does not accept input from the [ ] and [ ] keys.
Figure 6.2
Density / Calorie Conversion Factor Setting Mode
IM 11T03E01-01E
<6. CALORIE / PARAMETER SETTING>
6-10
(3) Display Mode
Select one of the following seven parameters to be shown in the message field. Each press of
the [NO] key cycles through the choices in the order shown below:
1. Physical density [XX.XXXX KG/M3] or [XX.XXXX LB/FT3]
(depends on the setting of CODE 21)
2. Compensated density [X.XXXX KG/M3 ]or [X.XXXXX LB/FT3]
(depends on the setting of CODE 21)
3. Calorific value [XXX.XXX MJ/M3] or [XX.XXXX KBTU]
(depends on the setting of CODE 22)
4. Temperature [XXX.X °C] or [XXX.X °F]
(depends on the setting of CODE 23)
5. Pressure [XXX.XX KPA] or [X.XXXX MPA] or [XX.XXX PSI]
(depends on the setting of CODE 20)
6. Current for analog output 1 (%) [XXX.X MA1%]
7. Current for analog output 2 (%) [XXX.X MA 2%]
Show the desired choice and press the [YES] key. That choice is set as a parameter to be
measured in the measurement mode (). Parameters 1, 2 and 3 are configurable,
while parameters 4 to 7 are read-only.
IM 11T03E01-01E
*1
<6. CALORIE / PARAMETER SETTING>
Measurement Mode
[MODE]
6-11
DISP
[YES]
XX.XXXX [YES] * 2, *3
KG/M3
(Physical density)
[NO]
2, 3
X.XXXX [YES] * *
KG/M3
(Compensated density)
[NO]
2, 3
XXX.XXX [YES] * *
MJ/M3
(Calorific value)
[NO]
2
[YES] *
XXX.X
(Read-only)
°C
(Sample gas temperature)
[NO]
2
XXX.XX [YES] *
(Read-only)
KPA
(Sample gas pressure)
[NO]
[YES]
XXX.X
(Read-only)
MA1%
(Current for analog output1(%))
[NO]
[YES]
XXX.X
(Read-only)
MA2%
(Current for analog output 2(%))
[NO]
Measurement Mode
F0603.ai
*1: See "Figure 6.1 Modes of Operation Level" on page 6-8.
*2: The unit is the one set the service level.
*3: Press the [YES] key to show the parameter in the measurement mode.
Figure 6.3
Display Mode (DISP Mode)
IM 11T03E01-01E
6.2.2
<6. CALORIE / PARAMETER SETTING>
6-12
Parameter Setting at Setting Level
To move to the setting level, press the [*] key and then enter the password (XXX). The main task
at the setting level is to set data values such as the measuring range values.
Note that the function for which you have set data values at the setting level, will not work if
that function is turned off at another level. For this reason, care must be taken when dealing
with modes that relate to each other. The setting level has four levels/modes, as shown below,
although this subsection explains only the analog output setting and alarm-point setting modes.
The calibration parameter setting mode is discussed in Chapter 9.
1. Analog output setting mode [*RANGE]
2. Calibration parameter setting mode [*CAL.DT]
3. Alarm-point setting mode [*ALARM]
4. Service level [*SERVC]
Select a mode from the above four choices. Each press of the [NO] key cycles through the
choices in sequence. Press the [YES] key to enter the desired mode when it is indicated.
Measurement Mode
*
[ ]
XXX
*PASSW
Password is not selected in first condition.
(See "Figure 6.28 Setting of Password" on page 6-31.)
[ENT]
*RANGE
[NO]
[YES]
See the process in Figure 6.5.
*CAL.DT
[NO]
[YES]
*ALARM
[NO]
[YES]
Calibration parameter setting mode
See the process in Figure 9.2.
Alarm-point setting mode
See the process in Figure 6.6.
*SERVC
[NO]
Analog output setting mode
[YES]
Service level
See the process in Figure 6.7 to 6.31.
[ENT]
[ENT]
[ENT]
[ENT]
F0604.ai
You can abort the setting operation in any of the modes by pressing the [MODE] key.
Doing this will return the analyzer to the measurement mode.
Figure 6.4
Process at the Setting Level
IM 11T03E01-01E
6-13
<6. CALORIE / PARAMETER SETTING>
1. Analog Output Setting Mode [*RANGE]
The analog output has output 1 and output 2. Only output 1 can be used for communication
purposes.
Set a range appropriate for 4-20 mA DC output signals. The range has two setpoints: the lower
limit (zero point) of the range corresponding to the minimum (0%) of a given output signal and the
upper limit (span point) of the range corresponding to the maximum (100%) of the given output
signal.
(1) Output 1 [*OUT1]
(a) Select a parameter to be output.
Physical density [*DENS]
Compensated density [*C_DNS]
Calorific value [*CALRY]
Gas temperature [*TEMP]
Gas pressure [*PRESS]
(b) Set the zero and span points. (This item is not applied if temperature or pressure are
chosen as the parameter in item (a) above.)
Physical density [*Z_DNS], [*S_DNS]
Configure range:
00.0000 to 60.0000 (kg/m3),
0.00000 to 4.00000 (lb/ft3)
Compensated density [*Z_CP.D], [*S_CP.D]
Configure range:
0.0000 to 6.0000 (kg/m3),
0.00000 to 0.40000 (lb/ft3)
Calorific value [*Z_CAL], [*S_CAL]
Configure range:
000.000 to 133.000(MJ/m3),
0.0000 to 3.5000 (kBTU/ft3)
(2) Output 2 [*OUT2]
The same parameters as those of item (1) apply.
IM 11T03E01-01E
<6. CALORIE / PARAMETER SETTING>
*1
6-14
Measurement mode
XXX
[*]
*PASSW
[ENT]
*SERVC
[NO]
*RANGE
[YES]
*OUT1
[YES]
*2
*DENS
(Physical density Range)
[NO]
*3
[NO]
XX.XXXX
[YES]
*Z_DNS
*CAL.DT
*OUT2
[YES]
XX.XXXX
[ENT]
*S_DNS
[ENT]
[NO]
*C_DNS
(Compensated
density Range)
[YES]
[NO]
*CALRY
(Calorific value Range)
X.XXXX
*Z_CP.D
X.XXXX
[ENT]
*Z_CAL
[ENT]
XXX.XXX
XXX.XXX
[YES]
*S_CP.D
[ENT]
*S_CAL
[ENT]
[NO]
*TEMP
(Gas temperature)
[YES]
[NO]
*PRESS
(Gas pressure)
[YES]
[NO]
F0605.ai
*1: You can abort the setting operation in any of the modes by pressing [MODE] key.
Doing this will return the analyzer to the measurement mode.
*2: The display shows the previous parameter first.
*3: Each press of the [NO] key toggles between the parameters [*OUT1] and [*OUT2].
Figure 6.5
Process of the Analog Output Setting Mode
IM 11T03E01-01E
6-15
<6. CALORIE / PARAMETER SETTING>
2. Alarm-point Setting Mode [*ALARM]
Set the upper /lower limits for the measured values of the parameters noted below, for the
purpose of raising an alarm.
(a) Select a parameter for which alarms are raised.
Physical density [*DENS]
Compensated density [*C_DNS]
Calorific value [*CALRY]
Press the [YES] key to enter the desired mode when it is indicated.
(b) Set the upper/lower limits.
Physical density [*L_DNS], [*H_DNS]
Configure range:
00.0000 to 60.0000 (kg/m3),
0.00000 to 4.00000 (lb/ft3)
Compensated density [*L_CP.D], [*H_CP.D]
Configure range:
0.0000 to 6.0000 (kg/m3),
0.00000 to 0.40000 (lb/ft3)
Calorific Value [*L_CAL], [*H_CAL]
Configure range:
000.000 to 133.000 (MJ/m3),
*1
*CAL.DT
[NO]
*ALARM
0.0000 to 3.5000 (kBTU/ft3)
[NO]
*SERVC
[YES]
*DENS
(Physical density)
XX.XXXX
[YES]
[NO]
*C_DNS
(Compensated density)
*L_DNS
XX.XXXX
[ENT]
X.XXXX
[YES]
*L_CP.D
*H_DNS
[ENT]
X.XXXX
[ENT]
*H_CP.D
[ENT]
[NO]
*2
*CALRY
(Calorific value)
XXX.XXX
[YES]
*L_CAL
XXX.XXX
[ENT]
*H_CAL
[ENT]
[NO]
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: The display shows the previous parameter first.
Figure 6.6
F0606.ai
Process of the Alarm-point Setting Mode
IM 11T03E01-01E
6.2.3
<6. CALORIE / PARAMETER SETTING>
6-16
Parameter Setting at Service Level
To move to the service level, select [*SERVC] at the setting level, and press the [YES] key.
(see Figure 6.4 on page 6-12.) The main task at the setting level is to set data values such as a
measuring range. To go into each mode of the service level, enter the appropriate code number.
CAUTION
DO NOT enter codes other than those listed in Table 6.4. A change to the settings of any
unlisted code can result in the failure of the analyzer to operate properly. If you have entered
a wrong code number by mistake, press the [MODE] key to return to the measurement
mode(). You may exit the mode of that wrong code by pressing the [ENT] key, if
you haven't made any change to the relevant data.
(1)
When [*SERVC] is shown on the display, press
[YES] key. The message field changes to look
like the one shown left.
(2)
Press the [ENT] key with code number 01 shown
on the display, for example. The analyzer enters
the mode for setting the parameters of that code.
0607.ai
Figure 6.7
Entry of Code Number at Service Level
The following explains the parameter setting procedures in the order of code numbers.
IM 11T03E01-01E
<6. CALORIE / PARAMETER SETTING>
6-17
CODE 01: Setting of Hold during Maintenance
This setting determines whether or not the output signal is held during maintenance. As a value
to be held, you can select either a value immediately before or a preset value.
(a) Hold enable/disable [*M_HLD]
Disable:
0
Enable:
1
Enable (preset value): 2
(b) Setting of preset hold value [*PR.SET]
This item applies if item (a) above is set to "2".
Configurable range:
-10.0 to 110.0 (percent of analog output)
*1
*SERVC
[YES]
01
*CODE
*2
[ENT]
X
*M_HLD
*5
*3
[ENT]
XXX.X
*PR.SET
*4
[ENT]
F0608.ai
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Set a value using the [ ] and [ENT] keys.
*4: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*5: Setting 0 or 1 returns to the service level with the indication [*SERVC].
Figure 6.8
Setting of Hold during Maintenance
IM 11T03E01-01E
<6. CALORIE / PARAMETER SETTING>
6-18
CODE 02: Setting of Hold in the Event of Errors
This setting determines whether or not the output signal is held if an error occurs. You can select
either the value immediately prior to the error or a preset value as the value to be held.
(a) Hold enable/disable [*E_HLD]
Disable:
0
Enable:
1
Enable (preset value):
2
(b) Setting of preset hold value [*PR.SET]
This item applies if item (a) above is set to "2".
Configurable range: -10.0 to 110.0 (percent of analog output)
*1
[YES]
*SERVC
02
*2
*CODE
[ENT]
X
*3
*E_HLD
*5
[ENT]
XXX.X
*PR.SET
*4
[ENT]
F0609.ai
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Set a value using the [ ] and [ENT] keys.
*4: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*5: Setting 0 or 1 returns to the service level with the indication [*SERVC].
Figure 6.9
Setting of Hold in the Event of Error
CODE 04: Setting of Output-smoothing Constants [*SMOTH]
This setting defines the constants for output smoothing.
This smoothing constants is output-smoothing constants of converter (electrical circuit boards).
Configurable range: 00 to 60 (unit : second)
*1
*SERVC
[YES]
04
*CODE
*2
[ENT]
XX
*SMOTH
*3
[ENT]
F0610.ai
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12 .
*2: Enter the code number.
*3: Set a value using the [ ] , [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
Figure 6.10
Setting of Output-smoothing Constants
IM 11T03E01-01E
<6. CALORIE / PARAMETER SETTING>
6-19
CODE 05: Setting of Contact Outputs Status [*CNTCT]
This setting defines the status of the contact outputs.
Configurable range: 00 to 15
*1
*SERVC
[YES]
05
*CODE
*2
[ENT]
XX
*CNTCT
*3
[ENT]
F0611.ai
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
Figure 6.11
Setting of Contact Outputs Status
Each contact output takes either of the following two status (in table); depending on the value you
set.
Value
ZERO contact
SPAN contact
MAINT contact
ALM contact
0
NO
NO
NO
NO
1
NO
NO
NO
NC
2
NO
NO
NC
NO
3
NO
NO
NC
NC
4
NO
NC
NO
NO
5
NO
NC
NO
NC
6
NO
NC
NC
NO
7
NO
NC
NC
NC
8
NC
NO
NO
NO
9
NC
NO
NO
NC
10
NC
NO
NC
NO
11
NC
NO
NC
NC
12
NC
NC
NO
NO
13
NC
NC
NO
NC
14
NC
NC
NC
NO
15
NC
NC
NC
NC
NO: means “normally open”
NC: means “normally closed”
IM 11T03E01-01E
<6. CALORIE / PARAMETER SETTING>
6-20
CODE 10: Setting of Pressure Compensation
This setting determines whether or not the measured density should be compensated by the gas
pressure. The density can be compensated to either the value immediately before the error or a
fixed value.
(a) Enable/disable/select preset value [*P.COMP]
Disable: 0
Enable (value immediately before): 1
Enable (fixed value):
2
(b) Setting of fixed value for pressure compensation [*P.FIX]
This item applies if item (a) above is set to "2".
Configurable range: 0.10 to 999.99 (for kPa),
0.0001 to 9.9999 (for MPa),
0.015 to 99.999 (for psi )
*1
*SERVC
[YES]
10
*CODE
*2
[ENT]
X
*P.COMP
*5
*3
[ENT]
XXX.XX
*P.FIX
*6
*4
[ENT]
F0612.ai
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Set a value using the [ ] and [ENT] keys.
*4: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*5: Setting 0 or 1 returns to the service level with the indication [*SERVC].
*6: The position of the decimal point changes depending on the setting of CODE 20.
If the code is set to "0" (kPa), the reading is XXX.XX;
if the code is set to "1" (MPa), the reading is X.XXXX;
if the code is set to "2" (psi), the reading is XX.XXX.
Figure 6.12
Setting of Pressure Compensation
IM 11T03E01-01E
<6. CALORIE / PARAMETER SETTING>
6-21
CODE 11: Setting for Compensated Density Measurement
This setting defines the reference temperature and pressure required to obtain the compensated
density.
(a) Setting of reference temperature [*C.D.TMP]
Configurable range:
-20.0 to 80.0 (for ºC)
-4.0 to 176.0 (for °F )
(b) Setting of reference pressure [*C.D.PRS]
Configurable range:
*1
*SERVC
[YES]
0.10 to 999.99 (for kPa),
0.0001 to 9.9999 (for MPa),
0.015 to 99.999 (for psi )
11
*CODE
*2
[ENT]
XX.X
*C.D.TMP
*4
*3
[ENT]
XXX.XX
*C.D.PRS
*5
*3
[ENT]
F0613.ai
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*4: The position of the decimal point changes depending on the setting of CODE 23.
If the code is set to "0" (°C), the reading is XX.X;
if the code is set to "1" (°F), the reading is XXX.X.
*5: The position of the decimal point changes depending on the setting of CODE 20.
If the code is set to "0" (kPa), the reading is XXX.XX;
if the code is set to "1" (MPa), the reading is X.XXXX;
if the code is set to "2" (psi), the reading is XX.XXX.
Figure 6.13
Setting for Compensated Density Measurement
IM 11T03E01-01E
<6. CALORIE / PARAMETER SETTING>
6-22
CODE 12: Setting of Gas Pressure
This setting defines the zero and span points of the pressure transmitter's measurement range.
(a) Setting of zero-point gas pressure [*Z_PRS]
Configurable range:
0.01 to 999.99 (for kPa),
0.0001 to 9.9999 (for MPa),
0.015 to 99.999 (for psi )
(b) Setting of span-point gas pressure [*S_PRS]
Configurable range: 0.10 to 999.99 (for kPa),
0.0001 to 9.9999 (for MPa),
0.015 to 99.999 (for psi )
The zero and span points set here are also used for the analog output.
*1
*SERVC
[YES]
12
*CODE
*2
[ENT]
XXX.XX
*Z_PRS
*4
*3
[ENT]
XXX.XX
*S_PRS
*4
*3
[ENT]
F0614.ai
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*4: The position of the decimal point changes depending on the setting of CODE 20.
If the code is set to "0" (kPa), the reading is XXX.XX;
if the code is set to "1" (MPa), the reading is X.XXXX;
if the code is set to "2" (psi), the reading is XX.XXX.
Figure 6.14
Setting of Gas Pressure
IM 11T03E01-01E
6-23
<6. CALORIE / PARAMETER SETTING>
CODE 13: Selection for Automatic Calibration [*AUTO.C]
This setting determines whether or not automatic calibration is carried out. If automatic calibration
is to be carried out, you must also configure CODE 15 at the service level.
Disable:
0
Enable:
1
*1
*SERVC
[YES]
13
*CODE
*2
[ENT]
X
*AUTO.C
*3
[ENT]
F0615
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
Figure 6.15
Selection for Automatic Calibration
CODE 14: Setting for Remote Semi-automatic Calibration [*REMOT]
This setting determines whether or not semi-automatic calibration is carried out remotely.
If remote semi-automatic calibration is to be carried out, you must wire the contact input.
You must also configure CODE 15 at the service level.
Disable:
0
Enable:
1
*1
*SERVC
[YES]
14
*CODE
*2
[ENT]
X
*REMOT
*3
[ENT]
F0616.ai
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
Figure 6.16
Setting for Remote Semi-automatic Calibration
IM 11T03E01-01E
<6. CALORIE / PARAMETER SETTING>
6-24
CODE 15: Setting of Data for Automatic/Semi-automatic Calibration
This setting defines the data needed to carry out automatic/semi-automatic calibration.
(a) Selection of calibration item [*CAL.P]
Zero and span:
0
Zero:
1
Span:
2
(b) Setting of calibration time [*CAL.T]
Configurable range: 00 to 59 (unit : minute)
(c) Setting of stabilization time [*STAB.T]
Configurable range: 00 to 59 (unit : minute)
(d) Setting of calibration starting time
(1) Setting of year/month/day [*Y_M_D]
Configurable range: 00.01.01 to 99.12.31
(2) Setting of hour/minute [*H_M]
Configurable range: 00.00 to 23.59
(e) Selection of unit of calibration interval [*CYCL.U]
Hour: 0
Day: 1
(f) Selection of calibration interval [*CYCL.T]
Item (e) above is set to "0".
Configurable range: 00 to 23 (for hour)
Item (e) above is set to "1".
Configurable range: 000 to 255 (for day)
For more information, see Chapter 9, “Calibration Procedure”.
IM 11T03E01-01E
6-25
<6. CALORIE / PARAMETER SETTING>
*1
*SERVC
[YES]
15
*CODE
*2
[ENT]
X
*CAL.P
*2
(Selection of calibration item)
[ENT]
XX
*CAL.T
*2
(Setting of calibration time)
[ENT]
XX
*STAB.T
*2
(Setting of stabilization time)
[ENT]
XX.XX.XX
*Y_M_D *2
(Setting of calibration starting time [year/month/day] )
[ENT]
XX.XX
*H_M
*2
(Setting of calibration starting time [hour/minute] )
[ENT]
X
*CYCL.U
*2
(Selection of unit of calibration interval)
[ENT]
XXX
*CYCL.T
*3
*2
(Selection of calibration interval)
[ENT]
F0617.ai
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*3: The number of digits on LCD varies depending on the unit of the calibration interval (parameter *CYCL.U).
If the parameter is set to "0" (hour), the indication is two-digit number.
If the parameter is set to "1" (day), the indication is three-digit number.
Figure 6.17
Setting of Data for Automatic/Semi-automatic Calibration
IM 11T03E01-01E
<6. CALORIE / PARAMETER SETTING>
6-26
CODE 20: Selection of Pressure Unit [*PRES.U]
This setting defines the pressure unit for the gas pressure transmitter.
kPa:
0
MPa: 1
psi :
2
*1
*SERVC
[YES]
20
*CODE
*2
[ENT]
X
*PRES.U
*3
[ENT]
F0618.ai
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
Figure 6.18
Selection of Pressure Unit
CODE 21: Selection of Density Unit [*DENS.U]
This setting defines the density unit for the gas density meter.
kg/m3: 0
lb/ft3:
1
*1
[YES]
*SERVC
21
*CODE
*2
[ENT]
X
*DENS.U
*3
[ENT]
F0619.ai
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
Figure 6.19
Selection of Density Unit
IM 11T03E01-01E
<6. CALORIE / PARAMETER SETTING>
6-27
CODE 22: Selection of Calorific Value Unit [ *CAL.U]
This setting defines the calorific value unit for the gas density meter.
MJ/m3 :
0
kBTU/ft3 :
*1
*SERVC
[YES]
1
22
*CODE
*2
[ENT]
X
*CAL.U
*3
[ENT]
F0620.ai
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
Figure 6.20
Selection of Calorific Value Unit
CODE 23: Selection of Temperature Unit [*TEMP.U]
This setting defines the temperature unit for the gas density meter.
°C:
0
°F:
1
*1
*SERVC
[YES]
23
*CODE
*2
[ENT]
X
*TEMP.U
*3
[ENT]
F0621
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
Figure 6.21
Selection of Temperature Unit
IM 11T03E01-01E
6-28
<6. CALORIE / PARAMETER SETTING>
CODE 30: Setting of Calendar
This setting defines the date and time.
(a) Setting of year/month/day [*Y_M_D]
Configurable range : 00.01.01 to 99.12.31
(b) Setting of hour/minute [*H_M]
Configurable range : 00.00 to 23.59
*1
*SERVC
[YES]
30
*CODE
*2
[ENT]
XX.XX.XX
*Y_M_D
*3
[ENT]
XX.XX
*H_M
*3
[ENT]
F0622.ai
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
Figure 6.22
Setting of Calendar
CODE 31: Showing/Hiding of Negative Measured Values [*MINUS]
This setting determines whether negative measured values (-) should be shown or hidden.
Show: 0
Hide: 1
*1
*SERVC
[YES]
31
*CODE
*2
[ENT]
X
*MINUS
*3
[ENT]
F0623.ai
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
Figure 6.23
Showing/Hiding of Negative Measured Values
IM 11T03E01-01E
6-29
<6. CALORIE / PARAMETER SETTING>
CODE 40: Indication of Calibration Coefficients (Read-only)
This setting determines the calibration coefficients to be indicated.
(a) Zero [*C_K_Z]:
indicated as SX.XXXX
(b) Span [*C_K_S]:
indicated as X.XXXX
*1
*SERVC
[YES]
40
*2
*CODE
[ENT]
SX.XXXX
*C_K_Z
*3
[ENT]
X.XXXX
*C_K_S
[ENT]
F0624.ai
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: By pressing the [ENT] key, the next indication is displayed.
This parameter does not accept the [ ] and [ ] key input.
Figure 6.24
Indication of Calibration Coefficients
CODE 41: Indication of Oscillation Frequency (Read-only)
This setting determines the F2 and F4 oscillation frequencies coefficients to be indicated.
(a) Indication of frequency F2 (kHz) [*F2.KHZ]:
indicated as X.XXXXX
(b) Indication of frequency F4 (kHz) [*F4.KHZ]:
indicated as X.XXXXX
(c) Indication of frequency ratio (kHz) [*F2/F4]:
indicated as X.XXXXX
*1
*SERVC
[YES]
41
*CODE
*2
[ENT]
X.XXXXX
*F2.KHZ
*3
[ENT]
X.XXXXX
*F4.KHZ
*3
[ENT]
X.XXXXX
*F2/F4
*3
[ENT]
F0625.ai
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: By pressing the [ENT] key, the next indication is displayed.
This parameter does not accept the [ ] and [ ] key input.
Figure 6.25
Indication of Oscillation Frequency
IM 11T03E01-01E
<6. CALORIE / PARAMETER SETTING>
6-30
CODE 42: Indication of Software Version [*REV] (Read-only)
This setting allows the software version to be confirmed (indicated).
Indicated as X.XX
*1
*SERVC
[YES]
42
*CODE
*2
[ENT]
X.XX
*REV
*3
[ENT]
F0626.ai
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Pressing the [ENT] key returns to the service level with the indication [*SERVC].
This parameter does not accept the [ ] and [ ] key input.
Figure 6.26
Indication of Software Version
CODE 43: Setting of High-resolution Mode [*S_CYC]
This setting selects between the normal mode and the high-resolution mode.
Normal mode:
0
High-resolution mode: 1
*1
*SERVC
[YES]
43
*CODE
*2
[ENT]
X
*S_CYC
*3
[ENT]
F0627.ai
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Set a value by using the [ ], [ ] and [ENT] keys.
Figure 6.27
Setting of High-resolution Mode
IM 11T03E01-01E
<6. CALORIE / PARAMETER SETTING>
6-31
CODE 44: Setting of PASSWORD [ *PASS]
This setting defines the password.
*1
*SERVC
[YES]
44
*CODE
*2
[ENT]
X.X.X
*PASS
*3
[ENT]
F0628.ai
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Set a value by using the [ ], [ ] and [ENT] keys.
X.X.X
Service level
Setting level
Operation level
Note: Password entry request ( if a password is set)
At the operation level, a request is issued when the
[MODE] key is pressed in the measurement mode. At the
setting level, it is issued when the [ * ] key is pressed. At
the service level, it is issued when the [YES] key is
pressed with [*SERVC] displayed.
Figure 6.28
Password
X
(Setting)
Contents
(0)
1
2
3
4
5
6
7
8
9
(No setting)
111
333
777
888
123
957
331
546
847
Setting of Password
CODE 45: Selecting Battery alarm detection [ *BAT]
This setting determines whether or not battery alarm is detection.
Non detection: 0
Detection:
1
*1
*SERVC
[YES]
45
*CODE
*2
[ENT]
X
*BAT
*3
[ENT]
F 629.ai
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Set a value by using the [ ], [ ] and [ENT] keys.
Figure 6.29
Selecting Battery alarm detection
IM 11T03E01-01E
<6. CALORIE / PARAMETER SETTING>
6-32
CODE 50: Selecting the instruments [ *MODEL]
This setting defines the model.
Density meter:
Calorie meter:
Hydrogen purity, replacement meter :
*1
*SERVC
[YES]
50
*CODE
0
1
2
*2
[ENT]
X
*MODEL
*3
[ENT]
F0630.ai
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Set a value by using the [ ], [ ] and [ENT] keys.
Figure 6.30
Selecting the instruments
CODE 82: Setting Detector Constant
This setting defines the detector constant.
CAUTION
The instruments may not only work improperly but also big problems occur unless the detector
constants entered incorrectly. Do carefully and precisely when enter the detector constants into
converter.
1. The GD402 gas density meter is shipped after adjusting the detector and converter in pairs.
When installation, confirm the converter serial number described on the label of detector so
that combines converter and detector correctly. If mismatched in pairs, converter may be out
of order. If combined correctly, no need to enter the detector constants again, as converter
has been adjusted with the constants in factory.
2. When supply converter or detector individually, enter the detector constants, described on
inside the lid of the GD40, into the converter so that the GD402 is going to be well.
IM 11T03E01-01E
<6. CALORIE / PARAMETER SETTING>
[YES]
*1
*SERVC
82
*CODE
6-33
*2
[ENT]
020
*PASSW
*4
[ENT]
XXXX
*K_A_H
*3
[ENT]
XXXX
*K_A_L
*3
[ENT]
XXXX
*K_B_H
*3
[ENT]
XXXX
*T2L_H
*3
[ENT]
XXXX
*T2L_L
*3
[ENT]
XXXX
*T3L_H
*3
[ENT]
XXXX
*T3L_L
*3
[ENT]
XXXX
*3
*K_B_L
[ENT]
XXXX
*3
*T1H_H
[ENT]
XXXX
*K_C_H
XXXX
*T1H_L
*3
[ENT]
XXXX
*K_C_L
*3
[ENT]
XXXX
*K_AH.H
*3
[ENT]
XXXX
*K_AH.L
*3
[ENT]
XXXX
*K_BH.H
*3
[ENT]
XXXX
*K_BH.L
*3
[ENT]
XXXX
*K_CH.H
*3
[ENT]
XXXX
*K_CH.L
*3
[ENT]
XXXX
*T1L_H
*3
[ENT]
XXXX
*T1L_L
*3
[ENT]
*3
[ENT]
XXXX
*T2H_H
*3
[ENT]
XXXX
*T2H_L
*3
[ENT]
XXXX
*T3H_H
*3
[ENT]
XXXX
*T3H_L
*3
[ENT]
XX.XX
*M_T1
*3
[ENT]
XX.XX
*M_T2
*3
[ENT]
XX.XX
*C_T
*3
[ENT]
XX.XXX
*TMP_Z
*3
[ENT]
XX.XXX
*TMP_S
*3
[ENT]
F0631.ai
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
Detector Constants are described inside the lid of detector GD40.
*4: Set the PASSWORD.
Figure 6.31
Setting Detector Constant
IM 11T03E01-01E
Blank Page
7.
7-1
<7. HYDROGEN PURITY / PARAMETER SETTING>
HYDROGEN PURITY/PARAMETER
SETTING
CAUTION
Password is not selected in first condition. When password is necessary, refer to Section “4.2
Setting Lists”.
This chapter describes how to set parameters of Hydrogen purity meter.
If select Density meter, see Chapter 5.
If select Calorie meter, see Chapter 6.
Conversion Table
1 lb/ft3
1 kBTU/ft3
= 16.01847 kg/m3
= 37.25901232 MJ/m3
X degF
= (X-32)/1.8
1 psi
= 6.894756794 kPa
Density of Air = 1.2928 kg/m3
= 0.08070684 lb/ft3
IM 11T03E01-01E
7.1
7-2
<7. HYDROGEN PURITY / PARAMETER SETTING>
Setting Parameters
Subsections 7.1.1 to 7.1.4 show the setting parameters for each level.
7.1.1
Setting Parameters at Measurement Level
When turned on, the analyzer starts up in the measurement mode ().
Table 7.1
Setting Parameters at Measurement Level
Mode/Setting parameter
Measurement mode
Indication of measurement value
Physical density
Compensated density
Hydrogen purity
Indication of message
Unit
Error symbol and number
Alarm mark and number
Indication of HOLD and FAIL
HOLD
FAIL
Selection of measuring range
Hydrogen purity
Substituents-concentration
Substituents-concentration
7.1.2
Display
Data to be Set
(or Conditions)
Remarks
See Figure 7.2 for key operation.
KG/M3
LB/FT3
KG/M3
LB/FT3
H2_AIR
H2_CO2
A_CO2
Set the unit in CODE 21.
Set the unit in CODE 21.
The display shows the measuring range
selected in the measuring range selection
mode.
ERR.NO
ALM.NO
HOLD
FAIL
S_GAS
H2_AIR
H2_CO2
A_CO2
Select the item to be
measured with hydrogen
purity meter.
H2 in AIR
H2 in CO2
AIR in CO2.
The unit is the one set in the service level.
Appears if an error occurs.
Appears if an alarm occurs.
Appears even during other modes.
Appears during a HOLD state.
Appears if a failure occurs.
See Figure 7.1 for key operation.
Setting Parameters at Operation Level
Press the [MODE] key and enter the password (XXX) to gain access to this level (see Figure 7.1
for key operation).
Table 7.2
Setting Parameters at Operation Level
Mode/Setting parameter
DISP mode
Physical density
Compensated density
Hydrogen purity
Sample gas temperature
Sample gas pressure
Analog output 1 (%)
Analog output 2 (%)
Calibration mode
Manual calibration
Display
DISP
KG/M3
LB/FT3
KG/M3
LB/FT3
H2_AIR
H2_CO2
A_CO2
ºC
ºF
KPA
MPA
PSI
MA1%
MA2%
MAN.CAL
Data to be Set
(or Conditions)
Remarks
Set the unit in CODE 21.
See Figure 7.2 for key operation.
Selection of parameters to be measured in
the measurement mode.
Set the unit in CODE 21.
Set the unit in CODE 23.
Read-only
Set the unit in CODE 20.
Read-only
Read-only
Read-only
See Figure 10.2 for key operation.
IM 11T03E01-01E
7.1.3
7-3
<7. HYDROGEN PURITY / PARAMETER SETTING>
Setting Parameters at Setting Level
Press the [*] key and enter the password (XXX) to gain access to this level (see Figure 7.4 for key
operation).
Table 7.3
Setting Parameters at Setting Level
Mode/Setting parameter
Analog output setting mode
Output 1
Set a parameter to be output
Display
Data to be Set
(or Conditions)
Remarks
*RANGE
*OUT1
*DENS
*C_DNS
*H_A
*H_C_A
*TEMP
*PRESS
Physical density
Compensated density
Hydrogen purity
Substituents-concentration
Gas temperature
No setting is required if the parameter is
Gas pressure
temperature or pressure.
*Z_DNS,
*S_DNS
00.0000 to 60.0000 kg/m3
0.00000 to 4.00000 lb/ft3
Compensated density
*Z_CP.D,
*S_CP.D
0.0000 to 6.0000 kg/m3
0.00000 to 0.40000 lb/ft3
Hydrogen purity
*Z_H_A,
*S_H_A
*Z_H.C.A,
*S_H.C.A
0.00000 to 0.40000
Set the zero and span points
Physical density
Substituents-concentration
Output 2 (same as output 1)
Setting of calibration data
Output hold value during
calibration
Setting of preset value
(Only if the preceding parameter
is set to 2 (preset value))
Alarm-point setting mode
Select a parameter for which an
alarm is raised
Physical density
Compensated density
Hydrogen purity
Set the high/low limits
Physical density
See Figure 7.5 for key operation.
0.0 to 100.0
*OUT2
*CAL.DT
*C_HLD
*PR.SET
· Disable: 0
· Enable (value
immediately before) : 1
· Enable (preset value) : 2
-10.0 to 110.0
*ALARM
See Figure 10.1 for key operation.
See Figure 7.6 for key operation.
*DENS
*C_DNS
*H_A
*L_DNS,
*H_DNS
0.0000 to 6.0000 kg/m3
0.00000 to 4.00000 lb/ft3
Compensated density
*L_CP.D,
*H_CP.D
0.0000 to 6.0000 kg/m3
0.00000 to 0.40000 lb/ft3
Hydrogen purity
*L_H_A
0.0 to 100.0
IM 11T03E01-01E
7.1.4
7-4
<7. HYDROGEN PURITY / PARAMETER SETTING>
Setting Parameters at Service Level
Press the [*] key and enter the password (XXX) to select the service level. When you select this
level, the analyzer shows a Code No. promptly. Type the appropriate code number and press the
[ENT] key.
Table 7.4
Code
No.
Setting Parameters at Service Level
Item
01 Setting of hold during maintenance
Hold enable/disable
Setting of preset hold value
02 Setting of hold against errors
Hold enable/disable
Setting of preset hold value
03 Setting of hydrogen purity range
Hold enable/disable
Setting of the preset hold value
Data to be Set
(or Conditions)
Display
*M_HLD
*PR.SET
*E_HLD
*PR.SET
*H_HLD
· Disable : 0
· Enable (value immediately
before) : 1
· Enable (preset value) : 2
-10.0 to 110.0
· Disable : 0
· Enable (value immediately
before) : 1
· Enable (preset value) : 2
-10.0 to 110.0
*PR.SET
· Disable : 0
· Enable (value immediately
before) : 1
· Enable (preset value) : 2
-10.0 to 110.0
Remarks
See Figure 7.8 for key operation.
If “preset value” is selected in the
preceding parameter.
See Figure 7.9 for key operation.
If “preset value” is selected in the
preceding parameter.
See Figure 7.10 for key operation.
If “preset value” is selected in the
preceding parameter.
04 Setting of output-smoothing constants
*SMOTH
00 to 60
See Figure 7.11 for key operation.
05 Setting of contact outputs status
*CNTCT
00 to 15
See Figure 7.12 for key operation.
10 Setting of pressure compensation
Enable/disable/fixed value
*P.COMP
Setting of fixed value for pressure
correction
11 Setting for measurement
compensated density
Setting of reference temperature
Setting of reference pressure
*P.FIX
See Figure 7.13 for key operation.
· Disable: 0
· Enable (measured value) : 1
· Enable (fixed value) : 2
0.10 to 999.99 kPa, 0.0001 to
If “fixed value” is selected in the
9.9999 MPa, 0.015 to 99.999 psi preceding parameter.
See Figure 7.14 for key operation.
*C.D.TMP
*C.D.PRS
12 Setting of gas pressure
Setting of zero-gas pressure
*Z_PRS
Setting of span-gas pressure
*S_PRS
-20.0 to 80.0°C, -4.0 to 176.0°F
0.10 to 999.99 kPa, 0.0001 to
9.9999 MPa, 0.015 to 99.999 psi
0.01 to 999.99 kPa, 0.0001 to
9.9999 MPa, 0.015 to 99.999 psi
0.10 to 999.99 kPa, 0.0001 to
9.9999 MPa, 0.015 to 99.999 psi
See Figure 7.15 for key operation.
14 Setting for remote semi-automatic
calibration
*REMOT
· Disable : 0
· Enable : 1
Open contact : AIR in CO2
Closed contact : H2 in CO2
See Figure 7.16 for key operation.
20 Selection of pressure unit
*PRES.U
· kPa : 0
· MPa : 1
· psi : 2
See Figure 7.17 for key operation.
21 Selection of density unit
*DENS.U
· kg/m3 : 0
· lb/ft3 : 1
See Figure 7.18 for key operation.
23 Selection of temperature unit
*TEMP.U
· °C : 0
· °F : 1
See Figure 7.19 for key operation.
31 Showing/hiding of negative measured
values (-)
*MINUS
· Show : 0
· Hide : 1
See Figure 7.20 for key operation.
IM 11T03E01-01E
Table 7.4
Code
No.
7-5
<7. HYDROGEN PURITY / PARAMETER SETTING>
Setting Parameters at Service Level
Item
40 Indication of calibration coefficients
(read-only)
Zero
Span
41 Indication of oscillation frequency
(read-only)
Frequency F2
Frequency F4
Frequency ratio
42 Indication of software version
(read-only)
43 Setting of high-resolution mode
44 Setting of password
45 Selecting Battery alarm detection
50 Selecting the instruments
82 Setting Detector Constant
Data to be Set
(or Conditions)
Display
Remarks
See Figure 7.21 for key operation.
*C_K_Z
*C_K_S
SX.XXXX
X.XXXX
See Figure 7.22 for key operation.
*F2.KHZ
*F4.KHZ
*F2/F4
*REV
*S_CYC
*PASS
*BAT
*MODEL
X.XXXXX
X.XXXXX
X.XXXXX
X.XX
See Figure 7.23 for key operation.
· Normal resolution : 0
· High-resolution : 1
See Figure 7.24 for key operation.
0.0.0 to 9.9.9
See Figure 7.25 for key operation.
· Non detection : 0
· Detection :
1
See Figure 7.26 for key operation.
· Density meter : 0
· Calorie meter : 1
· Hydrogen purity, replacement
meter : 2
See Figure 7.27 for key operation.
See Figure 7.28 for key operation.
IM 11T03E01-01E
7.2
<7. HYDROGEN PURITY / PARAMETER SETTING>
7-6
Parameter Setting
[Aborting the Setting Operation]
Press the [MODE] key. Normally, the [MODE] key is used to move from the measurement
() mode to a mode at the operation level. It also used to return to the measurement
mode from other modes.
7.2.1
Parameter Setting at Measurement Level
1. Measuring Range Selection Mode
Operate the keys as shown below to switch between the hydrogen purity meter and replacement
(substituents-concentration) meter.
MEASURE
[MODE]
XXX
*PASSW
Password is not selected in first condition.
(See "Figure 7.25 Setting of Password"
on page 7-27.)
[ENT]
S_GAS
[NO]
*2
[YES]
H2
purity meter
XXX.X
H2_AIR
*1
[YES]
[NO]
Replacement
(H2 in CO2)
meter
XXX.X
H2_CO2
*1
[YES]
[NO]
Replacement
(Air in CO2)
meter
XXX.X
A_CO2
*1
[YES]
[NO]
MEASURE
* 12
*
Figure 7.1
F0701.ai
: Press the [YES] key to show the parameter in the measurement mode.
: See "Figure 7.3 Modes of Operation Level" on page 7-9.
Measuring Range Selection Mode
IM 11T03E01-01E
7.2.2
7-7
<7. HYDROGEN PURITY / PARAMETER SETTING>
Parameter Setting at Operation Level
To move to the operation level, press the [MODE] key and then enter the password (XXX).
Password is not selected in first condition. When password is necessary, refer to Section 4.2
“Setting Lists” and selected password what you need.
This level includes the manual calibration mode, which is discussed in Chapter 10.
(1) Operation Level
Select a mode from the following four choices of the operation level. Each press of the [NO] key
cycles through the choices in the order shown below:
1. Measuring range selection mode [S_GAS]
2. Display mode [DISP]
3. Manual calibration mode [MAN.CAL]
Press the [YES] key to enter the desired mode when it is indicated.
Measurement Mode
[MODE]
XXX
*PASSW
Password is not selected in first condition
(See "Figure 7.25 Process of Setting Password" on page 7-29.)
[ENT]
S_GAS
[YES]
[NO]
DISP
[YES]
[NO]
MAN.CAL
[YES]
[NO]
Measuring range selection mode
See the process in Figure 7.1.
DISP mode
See the process in Figure 7.3.
Manual calibration mode
See the process in Figure 10.2.
[YES]
[YES]
[YES]
F0702.ai
You can abort the setting operation in any of the modes by pressing [MODE] key.
Doing this will return the analyzer to the measurement mode.
Figure 7.2
Modes of Operation Level
IM 11T03E01-01E
<7. HYDROGEN PURITY / PARAMETER SETTING>
7-8
2. Display Mode
Select one of the following seven parameters to be shown in the message field. Each press of
the [NO] key cycles through the choices in the order shown below:
1. Physical density [XX.XXXX KG/M3] or [XX.XXXXX LB/FT3]
(depends on the setting of CODE 21)
2. Compensated density [X.XXXX KG/M3 ]or [X.XXXXX LB/FT3]
(depends on the setting of CODE 21)
3. Concentration [XXX.X VOL%]
4. Temperature [XXX.X °C] or [XXX.X °F]
(depends on the setting of CODE 23)
5. Pressure [XXX.XX KPA] or [X.XXXX MPA] or [XX.XXX PSI]
(depends on the setting of CODE 20)
6. Current for analog output 1 (%) [XXX.X MA1%]
7. Current for analog output 2(%) [XXX.X MA 2%]
Show the desired choice and press the [YES] key. That choice is set as a parameter to be
measured in the measurement mode (). Parameters 1, 2 and 3 are configurable,
while parameters 4 to 7 are read-only.
IM 11T03E01-01E
*1
<7. HYDROGEN PURITY / PARAMETER SETTING>
Measurement Mode
7-9
DISP
[YES]
XX.XXXX [YES] * 2,* 3
KG/M3
(Physical density)
[NO]
X.XXXX [YES] * 2,* 3
KG/M3
(Compensated density)
[NO]
XXX.X
H2_AIR
[YES] * 2,* 3
(Hydrogen purity)
[NO]
XXX.X
°C
[YES] * 2
(Read-only)
(Sample gas temperature)
[NO]
XXX.XX [YES] * 2
(Read-only)
KPA
(Sample gas pressure)
[NO]
[YES]
XXX.X
(Read-only)
MA1%
(Current for analog output 1(%))
[NO]
[YES]
XXX.X
(Read-only)
MA2%
(Current for analog output 2(%))
[NO]
Measurement Mode
F0703.ai
*1: See "Figure 7.2 Modes of Operation Level" on page 7-7.
*2: The unit is the one set the service level.
*3: Press the [YES] key to show the parameter in the measurement mode.
Figure 7.3
Display Mode (DISP Mode)
IM 11T03E01-01E
7.2.3
<7. HYDROGEN PURITY / PARAMETER SETTING>
7-10
Parameter Setting at Setting Level
To move to the setting level, press the [*] key and then enter the password (XXX). The main task
at the setting level is to set data values such as the measuring range values.
Note that the function for which you have set data values at the setting level, will not work if
that function is turned off at another level. For this reason, care must be taken when dealing
with modes that relate to each other. The setting level has four levels/modes, as shown below,
although this subsection explains only the analog output setting and alarm-point setting modes.
The calibration parameter setting mode is discussed in Chapter 10.
1. Analog output setting mode [*RANGE]
2. Calibration parameter setting mode [*CAL.DT]
3. Alarm-point setting mode [*ALARM]
4. Service level [*SERVC]
Select a mode from the above four choices. Each press of the [NO] key cycles through the
choices in sequence. Press the [YES] key to enter the desired mode when it is indicated.
Measurement Mode
*
[ ]
XXX
*PASSW
Password is not selected in first condition.
(See "Figure 7.25 Setting of Password" on page 7-27.)
[ENT]
*RANGE
[NO]
[YES]
[YES]
*ALARM
[NO]
[YES]
Calibration parameter setting mode
See the process in Figure 10.1.
Alarm-point setting mode
See the process in Figure 7.6.
*SERVC
[NO]
Service level
[YES]
[ENT]
[ENT]
[ENT]
See the process in Figure 7.7 to 7.28.
You can abort the setting operation in any of the modes by pressing the [MODE] key.
Doing this will return the analyzer to the measurement mode.
Figure 7.4
[ENT]
See the process in Figure 7.5.
*CAL.DT
[NO]
Analog output setting mode
F0704.ai
Process at the Setting Level
IM 11T03E01-01E
7-11
<7. HYDROGEN PURITY / PARAMETER SETTING>
1. Analog Output Setting Mode [ *RANGE ]
The analog output has output 1 and output 2. Only output 1 can be used for communication
purposes.
Set a range appropriate for 4-20 mA DC output signals. The range has two set-points: the lower
limit (zero point) of the range corresponding to the minimum (0%) of a given output signal and the
upper limit (span point) of the range corresponding to the maximum (100%) of the given output
signal.
(1) Output 1 [*OUT1]
(a) Select a parameter to be output.
Physical density [*DENS]
Compensated density [*C_DNS]
Hydrogen purity [*H_A] *1
Substituents-concentration [*H_C_A] *2
Gas temperature [*TEMP]
Gas pressure [*PRESS]
*1: Not selectable (i.e., not shown) if the analyzer is in the replacement (substituents-concentration) meter mode.
*2: Not selectable (i.e., not shown) if the analyzer is in the hydrogen purity meter mode.
(b) Set the zero and span points. (This item is not applied if temperature or pressure are
chosen as the parameter in item (a) above.)
Physical density [*Z_DNS], [*S_DNS]
Configure range:
00.0000 to 60.0000 (kg/m3),
0.00000 to 4.00000 (lb/ft3)
Compensated density [*Z_CP.D], [*S_CP.D]
Configure range:
0.0000 to 6.0000 (kg/m3),
0.00000 to 0.40000 (lb/ft3)
Hydrogen purity [*Z_H_A] , [*S_H_A]
Configure range:
0.0 to 100.0
Substituents-concentration [*Z_H.C.A] , [*S_H.C.A]
Configure range:
0.0 to 100.0
(2) Output 2 [*OUT2]
The same parameters as those of item (1) apply.
Note:If the same parameter is selected for both outputs 1 and 2, the zero/span point settings you made the last time apply when
the parameter is output.
Example: Assume you first select parameter [*H_A] for analog output [*OUT1] and set the zero/span points as 100.0/0.0, and
then select parameter [*H_A] for analog output [*OUT2] and set the zero/span points as 0.0/100.0. In that case, the
parameter is output with the 0.0/100.0 zero/span point settings for both outputs 1 and 2.
IM 11T03E01-01E
<7. HYDROGEN PURITY / PARAMETER SETTING>
7-12
Measurement mode
*1
XXX
[*]
*PASSW
[ENT]
*ALARM
[NO]
[NO]
*SERVC
*RANGE
[NO]
*CAL.DT
[NO]
[YES]
*OUT1
[YES]
*2
*DENS
*3
[NO]
*OUT2
[YES]
[YES]
XX.XXXX
*Z_DNS
[ENT]
XX.XXXX
*S_DNS
[ENT]
[YES]
X.XXXX
*Z_CP.D
[ENT]
X.XXXX
*S_CP.D
[ENT]
[YES]
XXX.X
*Z_H_A
[ENT]
XXX.X
*S_H_A
[ENT]
[YES]
XXX.X
*Z_H.C.A
[ENT]
XXX.X
*S_H.C.A
[ENT]
(Physical density Range)
[NO]
*C_DNS
(Compensated density Range)
[NO]
*H_A
(Hydrogen purity Range)
[NO]
*H_C_A
(Substituents-concentration)
[NO]
*TEMP
(Gas temperature)
[YES]
[NO]
*PRESS
(Gas pressure)
[YES]
[NO]
F0705.ai
*1: You can abort the setting operation in any of the modes by pressing [MODE] key.
Doing this will return the analyzer to the measurement mode.
*2: The display shows the previous parameter first.
*3: Each press of the [NO] key toggles between the parameters [*OUT1] and [*OUT2].
Figure 7.5
Process of the Analog Output Setting Mode
IM 11T03E01-01E
7-13
<7. HYDROGEN PURITY / PARAMETER SETTING>
2. Alarm-point Setting Mode [*ALARM]
Set the upper /lower limits for the measured values of the parameters noted below, for the
purpose of raising an alarm.
(a) Select a parameter for which alarms are raised.
Physical density [*DENS]
Compensated density [*C_DNS]
Hydrogen purity [*H_A]
Press the [YES] key to enter the desired mode when it is indicated.
(b) Set the high/low limits.
Physical density [*L_DNS], [*H_DNS]
Configure range:
00.0000 to 60.0000 (kg/m3),
0.00000 to 4.00000 (lb/ft3)
Compensated density [*L_CP.D], [*H_CP.D]
Configure range:
0.0000 to 6.0000 (kg/m3),
0.00000 to 0.40000 (lb/ft3)
Hydrogen purity [*L_H_A]
Configure range:
*1
*CAL.DT
[NO]
0.0 to 100.0
[NO]
*ALARM
*SERVC
[NO]
*RANGE
[YES]
*DENS
(Physical density)
XX.XXXX
[YES]
[NO]
*C_DNS
*2
XX.XXXX
[ENT]
X.XXXX
[YES]
(Compensated density)
*3
*L_DNS
*L_CP.D
*H_DNS
[ENT]
X.XXXX
[ENT]
*H_CP.D
[ENT]
[NO]
XXX.X
*H_A
(Hydrogen purity)
[YES]
*L_H_A
[ENT]
[NO]
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: The display shows the previous parameter first.
*3: This step is skipped when the replacement meter is selected.
Figure 7.6
F0706.ai
Process of the Alarm-point Setting Mode
IM 11T03E01-01E
7.2.4
<7. HYDROGEN PURITY / PARAMETER SETTING>
7-14
Parameter Setting at Service Level
To move to the service level, select [*SERVC] at the setting level, and press the [YES] key.
(see Figure 7.4 on page 7-10.) The main task at the setting level is to set data values such as a
measuring range. To go into each mode of the service level, enter the appropriate code number.
CAUTION
DO NOT enter codes other than those listed in Table 7.4. A change to the settings of any
unlisted code can result in the failure of the analyzer to operate properly. If you have entered
a wrong code number by mistake, press the [MODE] key to return to the measurement
mode(). You may exit the mode of that wrong code by pressing the [ENT] key, if
you haven't made any change to the relevant data.
(1)
(2)
When [*SERVC] is shown on the display, press
[YES] key. The message field changes to look
like the one shown left.
Press the [ENT] key with code number 01 shown
on the display, for example. The analyzer enters
the mode for setting the parameters of that code.
F0707.ai
Figure 7.7
Entry of Code Number at Service Level
The following explains the parameter setting procedures in the order of code numbers.
IM 11T03E01-01E
<7. HYDROGEN PURITY / PARAMETER SETTING>
7-15
CODE 01: Setting of Hold during Maintenance
This setting determines whether or not the output signal is held during maintenance. As a value
to be held, you can select either a value immediately before or a preset value.
(a) Hold enable/disable [*M_HLD]
Disable:
0
Enable:
1
Enable (preset value): 2
(b) Setting of preset hold value [*PR.SET]
This item applies if item (a) above is set to "2".
Configurable range:
-10.0 to 110.0 (percent of analog output)
*1
*SERVC
[YES]
01
*CODE
*2
[ENT]
X
*M_HLD
*5
*3
[ENT]
XXX.X
*PR.SET
*4
[ENT]
F0708.ai
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: Set a value using the [ ] and [ENT] keys.
*4: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*5: Setting 0 or 1 returns to the service level with the indication [*SERVC].
Figure 7.8
Setting of Hold for During Maintenance
IM 11T03E01-01E
<7. HYDROGEN PURITY / PARAMETER SETTING>
7-16
CODE 02: Setting of Hold in the Event of Errors
This setting determines whether or not the output signal is held if an error occurs. You can select
either the value immediately prior to the error or a preset value as the value to be held.
(a) Hold enable/disable [*E_HLD]
Disable:
0
Enable:
1
Enable (preset value):
2
(b) Setting of preset hold value [*PR.SET]
This item applies if item (a) above is set to "2".
Configurable range: -10.0 to 110.0 (percent of analog output)
*1
*SERVC
[YES]
02
*2
*CODE
[ENT]
X
*E_HLD
*5
*3
[ENT]
XXX.X
*PR.SET
*4
F0709.ai
[ENT]
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: Set a value using the [ ] and [ENT] keys.
*4: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*5: Setting 0 or 1 returns to the service level with the indication [*SERVC].
Figure 7.9
Setting of Hold in the Event of Error
CODE 03: Setting of Hold of Hydrogen Purity Range
This setting enables or disables the function for holding the output in the hydrogen purity range
(hydrogen purity/replacement (substituents-concentration) meters mode only). You can select
either the value immediately prior to the hold or a preset value as the value to be held.
(a) Hold enable/disable [*H_HLD]
Disable:
0
Enable:
1
Enable (preset value): 2
IM 11T03E01-01E
<7. HYDROGEN PURITY / PARAMETER SETTING>
7-17
(b) Setting of preset hold value [*PR.SET]
This item applies if item (a) above is set to “2”.
Configurable range: -10.0 to 110.0 (percent of analog output)
*1
[YES]
*SERVC
03
*CODE
*2
[ENT]
X
*H_HLD
*5
*3
[ENT]
XXX.X
*PR.SET
*4
[ENT]
F0710.ai
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: Set a value using the [ ] and [ENT] keys.
*4: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*5: Setting 0 or 1 returns to the service level with the indication [*SERVC].
Figure 7.10
Process of Setting the Hold of Hydrogen Purity Range
[Note]
Relationship of CODE 01 to 03 with Output Hold Function Applied to Parameter [*CAL.DT] at
Setting level.
Hydrogen purity/Replacement (Substituents-concentration) Meters Mode
Priority
Hydrogen purity Range is
Selected for Output
Status
Hydrogen purity Range is
Not Selected for Output
1
Error
The output value selected in CODE03 The output value selected in CODE 02
2
Calibration
The output value selected in CODE03 The output value selected for data item
[*C_HLD] under parameter [*CAL.DT]
3
Maintenance
The output value selected in CODE03 The output valve selected in CODE 01
4
Switch to replacement
meter mode
The output value selected in CODE03 The normal output condition applies
Example: If you select parameter [*H_A] for output 1 and parameter [*H_C_A] for output 2, then
the resulting status-by-status output values to be held are summarized as follows:
Output 1
Output 2
During error status
As per CODE 03
As per CODE02
During calibration status
As per CODE 03
As per data item [*C_HLD ]
under parameter [*CAL.DT]
IM 11T03E01-01E
<7. HYDROGEN PURITY / PARAMETER SETTING>
7-18
CODE 04: Setting of Output-smoothing Constants [*SMOTH]
This setting defines the constants for output smoothing.
This smoothing constants is output-smoothing constants of converter (electrical circuit boards).
Configurable range: 00 to 60. (unit : second)
*1
*SERVC
[YES]
04
*CODE
*2
[ENT]
XX
*SMOTH
*3
[ENT]
F0711.ai
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
Figure 7.11
Setting of Output-smoothing Constants
CODE 05: Setting of Contact Outputs Status [*CNTCT]
This setting defines the status of the contact outputs.
Configurable range: 00 to 15
*1
*SERVC
[YES]
05
*CODE
*2
[ENT]
XX
*CNTCT
*3
[ENT]
F0712.ai
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
Figure 7.12
Setting of Contact Outputs Status
IM 11T03E01-01E
<7. HYDROGEN PURITY / PARAMETER SETTING>
7-19
Each contact output takes either of the following two status (in table); depending on the value you
set.
Value
SEL GAS contact
FUNC contact
MAINT contact
ALM contact
0
NO
NO
NO
NO
1
NO
NO
NO
NC
2
NO
NO
NC
NO
3
NO
NO
NC
NC
4
NO
NC
NO
NO
5
NO
NC
NO
NC
6
NO
NC
NC
NO
7
NO
NC
NC
NC
8
NC
NO
NO
NO
9
NC
NO
NO
NC
10
NC
NO
NC
NO
11
NC
NO
NC
NC
12
NC
NC
NO
NO
13
NC
NC
NO
NC
14
NC
NC
NC
NO
15
NC
NC
NC
NC
NO: means “normally open”
NC: means “normally closed”
IM 11T03E01-01E
<7. HYDROGEN PURITY / PARAMETER SETTING>
7-20
CODE 10: Setting of Pressure Compensation
This setting determines whether or not the measured density should be compensated by the
gas pressure. The density can be compensated either by receiving a signal from a pressure
transmitter or by using a fixed pressure set-point.
If you want the analyzer to compensate the density using a pressure transmitter, you MUST
connect a pressure transmitter to the converter and enter the pressure input and zero and span
points with CODE 12 at the service level.
If you do not want the analyzer to compensate the density, the analyzer shows and outputs the
measured density equivalent to a value at an atmospheric pressure of unity (101.33 kPa ABS)..
(a) Enable/disable/select preset value [*P.COMP]
Disable: 0
Enable (value immediately before): 1
Enable (fixed value):
2
(b) Setting of fixed value for pressure compensation [*P.FIX]
This item applies if item (a) above is set to "2".
Configurable range: 0.10 to 999.99 (for kPa),
0.0001 to 9.9999 (for MPa),
0.015 to 99.999 (for psi )
*1
*SERVC
[YES]
10
*CODE
*2
[ENT]
X
*P.COMP
*5
*3
[ENT]
XXX.XX
*P.FIX
*6
*4
[ENT]
F0713.ai
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: Set a value using the [ ] and [ENT] keys.
*4: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*5: Setting 0 or 1 returns to the service level with the indication [*SERVC].
*6: The position of the decimal point changes depending on the setting of CODE 20.
If the code is set to "0" (kPa), the reading is XXX.XX;
if the code is set to "1" (MPa), the reading is X.XXXX;
if the code is set to "2" (psi), the reading is XX.XXX.
Figure 7.13
Setting of Pressure Compensation
IM 11T03E01-01E
<7. HYDROGEN PURITY / PARAMETER SETTING>
7-21
CODE 11: Setting for Compensated Density Measurement
This setting defines the reference temperature and pressure required to obtain the compensated
density.
(a) Setting of reference temperature [*C.D.TMP]
Configurable range:
-20.0 to 80.0 (for ºC)
-4.0 to 176.0 (for °F )
(b) Setting of reference pressure [*C.D.PRS]
Configurable range:
*1
*SERVC
[YES]
0.10 to 999.99 (for kPa),
0.0001 to 9.9999 (for MPa),
0.015 to 99.999 (for psi )
11
*CODE
*2
[ENT]
XX.X
*C.D.TMP
*4
*3
[ENT]
XXX.XX
*C.D.PRS
*5
*3
[ENT]
F0714.ai
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER.]
Press the [YES] or [NO] key to re-set the value.
*4: The position of the decimal point changes depending on the setting of CODE 23.
If the code is set to "0" (°C), the reading is XX.X;
if the code is set to "1" (°F), the reading is XXX.X.
*5: The position of the decimal point changes depending on the setting of CODE 20.
If the code is set to "0" (kPa), the reading is XXX.XX;
if the code is set to "1" (MPa), the reading is X.XXXX;
if the code is set to "2" (psi), the reading is XX.XXX.
Figure 7.14
Setting for Compensated Density Measurement
IM 11T03E01-01E
<7. HYDROGEN PURITY / PARAMETER SETTING>
7-22
CODE 12: Setting of Gas Pressure
This setting defines the zero and span points of the pressure transmitter's measurement range.
(a) Setting of zero-point gas pressure [*Z_PRS]
Configurable range:
0.01 to 999.99 (for kPa),
0.0001 to 9.9999 (for MPa),
0.015 to 99.999 (for psi )
(b) Setting of span-point gas pressure [*S_PRS]
Configurable range: 0.10 to 999.99 (for kPa),
0.0001 to 9.9999 (for MPa),
0.015 to 99.999 (for psi )
The zero and span points set here are also used for the analog output.
*1
*SERVC
[YES]
12
*CODE
*2
[ENT]
XXX.XX
*Z_PRS
*4
*3
[ENT]
XXX.XX
*S_PRS
*4
*3
[ENT]
F0715.ai
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*4: The position of the decimal point changes depending on the setting of CODE 20.
If the code is set to "0" (kPa), the reading is XXX.XX;
if the code is set to "1" (MPa), the reading is X.XXXX;
if the code is set to "2" (psi), the reading is XX.XXX.
Figure 7.15
Setting of Gas Pressure
IM 11T03E01-01E
<7. HYDROGEN PURITY / PARAMETER SETTING>
7-23
CODE 14: Setting for Remote Semi-automatic Calibration [*REMOT]
This setting determines whether or not the measuring range of the replacement (substituentsconcentration) meter is selected remotely.
If remote range selection is to be carried out, you must wire the contact input.
Disable:
0
Enable:
1
If the code is set to “1”, the analyzer selects:
the “AIR in CO2” range, if the contact input is open.
the “H2 in CO2” range, if the contact input is closed.
*1
*SERVC
[YES]
14
*CODE
*2
[ENT]
X
*REMOT
*3
[ENT]
F0716.ai
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
Figure 7.16
Setting for Remote Semi-automatic Calibration
CODE 20: Selection of Pressure Unit [*PRES.U]
This setting defines the pressure unit for the gas pressure transmitter.
kPa:
0
MPa: 1
psi :
2
*1
*SERVC
[YES]
20
*CODE
*2
[ENT]
X
*PRES.U
*3
[ENT]
F0717.ai
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
Figure 7.17
Selection of Pressure Unit
IM 11T03E01-01E
<7. HYDROGEN PURITY / PARAMETER SETTING>
7-24
CODE 21: Selection of Density Unit [*DENS.U]
This setting defines the density unit for the gas density meter.
kg/m3: 0
lb/ft3:
1
*1
[YES]
*SERVC
21
*CODE
*2
[ENT]
X
*DENS.U
*3
[ENT]
F0718.ai
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
Figure 7.18
Selection of Density Unit
CODE 23: Selection of Temperature Unit [*TEMP.U]
This setting defines the temperature unit for the gas density meter.
°C:
0
°F:
1
*1
*SERVC
[YES]
23
*CODE
*2
[ENT]
X
*TEMP.U
*3
[ENT]
F0719.ai
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
Figure 7.19
Selection of Temperature Unit
IM 11T03E01-01E
7-25
<7. HYDROGEN PURITY / PARAMETER SETTING>
CODE 31: Showing/Hiding of Negative Measured Values [*MINUS]
This setting determines whether negative measured values (-) should be shown or hidden.
Show: 0
Hide: 1
*1
*SERVC
[YES]
31
*2
*CODE
[ENT]
X
*3
*MINUS
[ENT]
F0720.ai
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
Figure 7.20
Showing/Hiding of Negative Measured Values
CODE 40: Indication of Calibration Coefficients (Read-only)
This setting determines the calibration coefficients to be indicated.
(a) Zero [*C_K_Z]:
indicated as SX.XXXX
(b) Span [*C_K_S]:
indicated as X.XXXX
*1
*SERVC
[YES]
40
*CODE
*2
[ENT]
SX.XXXX
*C_K_Z
*3
[ENT]
X.XXXX
*C_K_S
[ENT]
F0721.ai
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: By pressing the [ENT] key, the next indication is displayed.
This parameter does not accept the [ ] and [ ] key input.
Figure 7.21
Indication of Calibration Coefficients
IM 11T03E01-01E
7-26
<7. HYDROGEN PURITY / PARAMETER SETTING>
CODE 41: Indication of Oscillation Frequency (Read-only)
This setting determines the F2 and F4 oscillation frequencies coefficients to be indicated.
(a) Indication of frequency F2 (kHz) [*F2.KHZ]:
indicated as X.XXXXX
(b) Indication of frequency F4 (kHz) [*F4.KHZ]:
indicated as X.XXXXX
(c) Indication of frequency ratio (kHz) [*F2/F4]:
indicated as X.XXXXX
*1
*SERVC
[YES]
41
*CODE
*2
[ENT]
X.XXXXX
*F2.KHZ
*3
[ENT]
X.XXXXX
*F4.KHZ
*3
[ENT]
X.XXXXX
*F2/F4
*3
[ENT]
F0722.ai
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: By pressing the [ENT] key, the next indication is displayed.
This parameter does not accept the [ ] and [ ] key input.
Figure 7.22
Indication of Oscillation Frequency
CODE 42: Indication of Software Version [*REV] (Read-only)
This setting allows the software version to be confirmed (indicated).
Indicated as X.XX
*1
*SERVC
[YES]
42
*CODE
*2
[ENT]
X.XX
*REV
*3
[ENT]
F0723.ai
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: Pressing the [ENT] key returns to the service level with the indication [*SERVC].
This parameter does not accept the [ ] and [ ] key input.
Figure 7.23
Indication of Software Version
IM 11T03E01-01E
<7. HYDROGEN PURITY / PARAMETER SETTING>
7-27
CODE 43: Setting of High-resolution Mode [*S_CYC]
This setting selects between the normal mode and the high-resolution mode.
Normal mode:
0
High-resolution mode: 1
*1
*SERVC
[YES]
43
*CODE
*2
[ENT]
X
*S_CYC
*3
[ENT]
F0724.ai
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: Set a value by using the [ ], [ ] and [ENT] keys.
Figure7.24
Setting of High-resolution Mode
CODE 44: Setting of PASSWORD [*PASS]
This setting defines the password.
*1
*SERVC
[YES]
44
*CODE
*2
[ENT]
X.X.X
*PASS
*3
[ENT]
F0725
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: Set a value by using the [ ], [ ] and [ENT] keys.
X.X.X
Service level
Setting level
Operation level
Note: Password entry request ( if a password is set)
At the operation level, a request is issued when the
[MODE] key is pressed in the measurement mode. At the
setting level, it is issued when the [ * ] key is pressed. At
the service level, it is issued when the [YES] key is
pressed with [*SERVC] displayed.
Figure 7.25
Password
X
(Setting)
Contents
(0)
1
2
3
4
5
6
7
8
9
(No setting)
111
333
777
888
123
957
331
546
847
Setting of Password
IM 11T03E01-01E
<7. HYDROGEN PURITY / PARAMETER SETTING>
7-28
CODE 45: Selecting Battery alarm detection [*BAT]
This setting determines whether or not battery alarm is detection.
Non detection: 0
Detection:
1
*1
*SERVC
[YES]
45
*CODE
*2
[ENT]
X
*BAT
*3
[ENT]
F0726
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: Set a value by using the [ ], [ ] and [ENT] keys.
Figure 7.26
Selecting Battery alarm detection
CODE 50: Selecting the instruments [*MODEL]
This setting defines the model.
Density meter:
Calorie meter:
Hydrogen purity, replacement meter :
*1
*SERVC
[YES]
50
*CODE
0
1
2
*2
[ENT]
X
*MODEL
*3
[ENT]
F0727
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: Set a value by using the [ ], [ ] and [ENT] keys.
Figure 7.27
Selecting the instruments
IM 11T03E01-01E
<7. HYDROGEN PURITY / PARAMETER SETTING>
7-29
CODE 82: Setting Detector Constant
This setting defines the detector constant.
CAUTION
The instruments may not only work improperly but also big problems occur unless the detector
constants entered incorrectly. Do carefully and precisely when enter the detector constants into
converter.
1. The GD402 gas density meter is shipped after adjusting the detector and converter in pairs.
When installation, confirm the converter serial number described on the label of detector so
that combines converter and detector correctly. If mismatched in pairs, converter may be out
of order. If combined correctly, no need to enter the detector constants again, as converter
has been adjusted with the constants in factory.
2. When supply converter or detector individually, enter the detector constants, described on
inside the lid of the GD40, into the converter so that the GD402 is going to be well.
IM 11T03E01-01E
<7. HYDROGEN PURITY / PARAMETER SETTING>
[YES]
*1
*SERVC
82
*CODE
7-30
*2
[ENT]
020
*PASSW
*4
[ENT]
XXXX
*K_A_H
*3
[ENT]
XXXX
*K_A_L
*3
[ENT]
XXXX
*K_B_H
*3
[ENT]
XXXX
*T2L_H
*3
[ENT]
XXXX
*T2L_L
*3
[ENT]
XXXX
*T3L_H
*3
[ENT]
XXXX
*T3L_L
*3
[ENT]
XXXX
*3
*K_B_L
[ENT]
XXXX
*3
*T1H_H
[ENT]
XXXX
*K_C_H
XXXX
*T1H_L
*3
[ENT]
XXXX
*K_C_L
*3
[ENT]
XXXX
*K_AH.H
*3
[ENT]
XXXX
*K_AH.L
*3
[ENT]
XXXX
*K_BH.H
*3
[ENT]
XXXX
*K_BH.L
*3
[ENT]
XXXX
*K_CH.H
*3
[ENT]
XXXX
*K_CH.L
*3
[ENT]
XXXX
*T1L_H
*3
[ENT]
XXXX
*T1L_L
*3
[ENT]
*3
[ENT]
XXXX
*T2H_H
*3
[ENT]
XXXX
*T2H_L
*3
[ENT]
XXXX
*T3H_H
*3
[ENT]
XXXX
*T3H_L
*3
[ENT]
XX.XX
*M_T1
*3
[ENT]
XX.XX * 3
*M_T2
[ENT]
XX.XX
*C_T
*3
[ENT]
XX.XXX
*TMP_Z
*3
[ENT]
XX.XXX
*TMP_S
*3
[ENT]
F0728
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
Detector Constants are described inside the lid of detector GD40.
*4: Set the PASSWORD.
Figure 7.28
Setting Detector Constant
IM 11T03E01-01E
8.
<8. DENSITY / CALIBRATION PROCEDURE>
8-1
DENSITY /CALIBRATION
PROCEDURE
This chapter describes how to calibrate of Density meter.
If select Calorie meter, see Chapter 9.
If select Hydrogen purity meter, see Chapter 10.
Zero and span calibration can be carried out using standard gases (zero and span gases). Three
modes are available for calibration: automatic, semi-automatic (remote semi-automatic) and
manual. Even if automatic calibration is enabled in the parameter setting, you can interrupt the
system with either the semi-automatic or manual mode of calibration. The following explains the
basic procedure and operation used in calibration.
8.1
Basic Calibration Procedure
Calibration parameter setting
See the process in Figure 8.2
Selection of semi-automatic/manual calibration
See the process in Figure 5.1
Manual calibration mode
See the process in Figure 8.4
Semi-automatic calibration mode
See the process in Figure 8.3.
F0801.ai
8.2
Preparation for Calibration
Prepare for calibration during the initial operation of the analyzer, by carrying out the following.
At the operation level, set the parameters for semi-automatic and manual calibration.
Also set the valve-operation parameter for switching between gases to check the valve
performance.
• At the setting level, set the calibration data (zero- and span-point values, status of output
during calibration etc.)
• At the service level, determine whether automatic and remote semi-automatic calibrations
are enabled or disabled in CODE 13 and 14. Set the calibration data (selection of calibration
item, calibration time, stabilization time, etc.) in CODE 15.
8.2.1
Setting Type of Calibration and Checking Valve
Operation
(1) Operations to Set the Semi-automatic and Manual Calibration
In the measurement mode, press the [MODE] key, (if password is selected, enter the password
(XXX) (See section “4.2 Setting Lists”), ) and select the desired type of calibration.
• To select semi-automatic calibration, show [SEM.CAL] and press the [YES] key.
• To select manual calibration, show [MAN.CAL] and press the [YES] key.
For details on this process, see Figure 5.1.
IM 11T03E01-01E
<8. DENSITY / CALIBRATION PROCEDURE>
8-2
(2) Operations for Setting the Automatic/Remote Semi-automatic Calibration
• Selection for automatic calibration [*AUTO.C]
At the service level, select CODE 13.
• Disable: 0
• Enable: 1
• Selection for remote semi-automatic calibration [*REMOT]
At the service level, select CODE 14.
• Disable: 0
• Enable: 1
(3) Valve Check
Check the performance of valves according to the “valve operation flowchart” at the operation
level.
*1
MAN.CAL
[NO]
[NO]
VALVE
DISP
[YES]
*7
*7
XXX.XXX
V_ZERO
[YES]*4
XXX.XXX
Z_OPEN
[NO] *2
(Zero-gas valve operation)
[NO]
*7
*7
XXX.XXX
V_SPAN
XXX.XXX
[YES]*4
S_OPEN
[NO] *2
(Span-gas valve operation)
[NO]
[NO]*3
END
[YES]
WAIT
*6
(Stabilization time)
*5
[YES]
or
[NO]
F0802.ai
Measurement mode
*1: See "Figure 5.1 Modes of Operation Level" on page 5-8.
*2: Press the [NO] key to return to the parameter for sample gas.
*3: Using the [NO] key, show [V_ZERO], [V_SPAN] and [END] in succession.
*4: Using the [YES] key, open the zero-gas or span-gas value to read the measured value
at the point.
*5: Press the [YES] key to go into the measurement mode state after a given stabilization
time elapses; press the [NO] key to go into that state immediately.
*6: This step is skipped if no valve was opened.
*7: The display shows the same data item to be measured as the one in the measurement mode.
Figure 8.1
Valve Operation Mode
IM 11T03E01-01E
<8. DENSITY / CALIBRATION PROCEDURE>
8.2.2
8-3
Setting Calibration Data
To set calibration data, press the [*] key in the measurement mode, (when password is selected,
enter the password (XXX), ) and select the parameter [*CAL.DT]. See Figure 5.4 for the process
at the setting level.
(a) Zero-point density [*Z_DNS]
Configurable range : 0.0000 to 6.0000 (kg/m3)
0.00000 to 0.40000 (lb/ft3)
(b) Span-point density [*S_DNS]
Configurable range :
0.0000 to 6.0000 (kg/m3)
0.00000 to 0.40000 (lb/ft3)
(c) Output hold value during calibration [*C_HLD]
• Disable:
0
• Enable:
1
• Enable (preset value):
2
(d) Setting of preset value [*PR.SET]
Configurable range:
-10.0 to 110.0
(Applies only if item (c) above is set to “2”.)
*1
*RANGE
[NO]
*CAL.DT
[NO]
*ALARM
[YES]
X.XXXX
*Z_DNS
*2
(Setting of Zero-point density)
[ENT]
X.XXXX
*S_DNS
*2
(Setting of Span-point density)
[ENT]
X
*C_HLD
*2
(Setting of Output hold value during calibration)
[ENT]
XXX.X
*PR.SET
*2
*3
(Setting of preset value)
F0803.ai
[ENT]
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*3: This step is skipped if the parameter [*C_HLD] is set to a value other than "2".
Figure 8.2
Calibration Parameter Setting Mode
IM 11T03E01-01E
<8. DENSITY / CALIBRATION PROCEDURE>
8-4
8.3 Calibration
8.3.1
Semi-automatic Calibration
In this mode of calibration, you can carry out one-touch calibration. To use this feature, you must
observe the following instructions.
• The analyzer needs solenoid valves for controlling the calibration gases (zero and span
gases). Connect the solenoid valves with utmost care to ensure that there is no gas leakage
in the system.
• Install the calibration gas cylinders in a location close to the GD40 detector. If the detector
needs to be installed in an explosion-hazardous area, the solenoid valves being used must
also be protected against explosion.
• The contacts for controlling solenoid valves are built in the GD402T, GD402V, GD402R
converter. It is the user's responsibility to supply power to drive the contacts. You must install
a switch and a fuse along the power line. Note that the contacts are rated at 250 V AC/3 A or
30 V DC/3 A.
*1
DISP
[NO]
[NO]
SEM.CAL
MAN.CAL
[YES]
[NO]
START
[YES]
XX.XXXX
ZERO
XX.XXXX
SPAN
XX.XXXX
WAIT
* 34
*5
*
Zero-point density display
*3
*5
Span-point density display
*2
*3
Stabilization time
Measurement mode
*2
F0804.ai
*1: See "Figure 5.1 Modes of Operation Level" on page 5-8.
*2: If a calibration data error [ALM.10] occurs, the display alternates between the [WAIT] and [ALM.10] indications.
The [ALM.10] indication still remains on display even after the mode move to the measurement mode.
This error massage does not disappear until normally ending in re-calibration or resetting the analyzer.
*3: The values are operated and the calibration data are read at the time setpoint (calibration time) which is defined
in the mode for setting automatics/semi-automatic calibration data in CODE 15.
The display always shows the corrected density.
*4: During automatic calibration, the display shows the same data items as those of the [ZERO] and subsequent
indications.
*5: When pressing the [MODE] key during [ZERO] or [SPAN] status, the [WAIT] appears on the display.
When pressing the [MODE] key once again, the analyzer returns to the measurement mode.
(This operation means you have cancelled automatic/semi-automatic calibration.)
Figure 8.3
Semi-automatic Calibration Mode
IM 11T03E01-01E
<8. DENSITY / CALIBRATION PROCEDURE>
8.3.2
8-5
Manual Calibration
In this mode of calibration, you visually make sure the reading has settled, and then confirm the
reading manually (by pressing the [ENT] key). There are two ways to switch between the zero,
span and sample gases: by using manually-operated valves or by using solenoid valves. When
switching between the gases using solenoid valves, follow the instructions on valve installation in
Subsection 8.3.1.
*1
SEM.CAL
[NO]
[NO]
MAN.CAL
[YES]
ZERO
VALVE
[NO]
[YES]
X.XXXX
Z_DNS
*3
[ENT]
Zero-point density display
X.XXXX
CAL.SET
*6
[NO]
*4
[ENT]
SPAN
[YES]
X.XXXX
S_DNS
Span-point density display
*3
[ENT]
X.XXXX
CAL.SET
*4
*2
[ENT]
ALM.10
X.XXXX
WAIT
*5
[YES] or [NO]
Measurement mode
F0805.ai
*1: See "Figure 5.1 Modes of Operational Level" on page 5-8.
*2: If a calibration data error occurs, the display shows [ALM.10].
Pressing the [YES] or [NO] key returns the display to the [MAN.CAL].
*3: Set a desired value by using the [ ], [ ] and [ENT] keys.
*4: When the reading settles, press the [ENT] key to confirm your calibration.
This parameter does not accept input with the [ ] and [ ] keys.
If you press the [MODE] key (for ESCAPE) after this parameter appears, the display shows [WAIT].
The display always shows the corrected density.
*5: Press the [YES] key to go into the measurement mode status after a given stabilization time elapses: Press the [NO]
key to go into the state immediately.
*6: In case of pressing the [NO] key after skipping zero calibration (by pressing [NO] key in the ZERO status), the
analyzer returns to [MAN.CAL].
In case of pressing the [NO] key after finishing zero calibration, the analyzer goes to [WAIT] status.
Figure 8.4
Manual Calibration Mode
IM 11T03E01-01E
8.3.3
<8. DENSITY / CALIBRATION PROCEDURE>
8-6
Automatic Calibration
This mode of calibration takes effect when you select “enable” for automatic calibration in CODE
13 - [*AUTO.C] - at the service level. Even during automatic calibration, you can interrupt the
system with the semi-automatic or manual mode of calibration. One cycle of automatic calibration
is skipped, however, if the time of automatic calibration arrives when semi-automatic calibration is
in progress.
Measurement mode
X.XXXX
ZERO
X.XXXX
SPAN
*1
*1
X.XXXX
WAIT
*2
Measurement mode
F0806.ai
*1 : Pressing the [MODE] key, change it to display [WAIT] indication.
*2 : Pressing the [MODE] key, once again returns it measuring mode.
Figure 8.5
Automatic Calibration Mode
IM 11T03E01-01E
<9. CALORIE / CALIBRATION PROCEDURE>
9.
9-1
CALORIE /CALIBRATION
PROCEDURE
This chapter describes how to calibrate of Calorie meter.
If select Density meter, see Chapter 8.
If select Hydrogen purity meter, see Chapter 10.
Zero and span calibration can be carried out using standard gases (zero and span gases). Three
modes are available for calibration: automatic, semi-automatic (remote semi-automatic) and
manual. Even if automatic calibration is enabled in the parameter setting, you can interrupt the
system with either the semi-automatic or manual mode of calibration. The following explains the
basic procedure and operation used in calibration.
9.1
Basic Calibration Procedure
Calibration parameter setting
See the process in Figure 9.2.
Density/calorie conversion factor setting
See the process in Figure 6.2.
Selection of semi-automatic/manual calibration
See the process in Figure 6.1.
Manual calibration mode
Semi-automatic calibration mode
See the process in Figure 9.4.
See the process in Figure 9.3.
F0901.ai
9.2
Preparation for Calibration
Prepare for calibration during the initial operation of the analyzer, by carrying out the following.
• At the operation level, set the parameters for semi-automatic and manual calibration.
Also set the valve-operation parameter for switching between gases to check the valve
performance.
• At the setting level, set the calibration data (zero- and span-point values, status of output
during calibration etc.)
• At the service level, determine whether automatic and remote semi-automatic calibrations
are enabled or disabled in CODE 13 and 14. Set the calibration data (selection of calibration
item, calibration time, stabilization time, etc.) in CODE 15.
9.2.1
Setting Type of Calibration and Checking Valve
Operation
(1) Operations to Set the Semi-automatic and Manual Calibration
In the measurement mode, press the [MODE] key, (if password is selected, enter the password
(XXX) (See section “4.2 Setting Lists”), ) and select the desired type of calibration.
• To select semi-automatic calibration, show [SEM.CAL] and press the [YES] key.
• To select manual calibration, show [MAN.CAL] and press the [YES] key.
For details on this process, see Figure 6.1.
IM 11T03E01-01E
9-2
<9. CALORIE / CALIBRATION PROCEDURE>
(2) Operations for Setting the Automatic/Remote Semi-automatic Calibration
• Selection for automatic calibration [*AUTO.C]
At the service level, select CODE 13.
• Disable:
0
• Enable:
1
• Selection for remote semi-automatic calibration [*REMOT]
At the service level, select CODE 14.
• Disable:
0
• Enable:
1
(3) Valve Check
Check the performance of valves according to the “valve operation flowchart” at the
operation level.
*1
MAN.CAL
[NO]
[NO]
VALVE
[YES]
*7
DISP
*7
XXX.XXX
XXX.XXX
V_ZERO
[YES] * 4
(Zero-gas valve operation)
Z_OPEN
[NO]* 2
[NO]
*7
*7
XXX.XXX
XXX.XXX
V_SPAN
[YES] * 4
(Span-gas valve operation)
S_OPEN
[NO] * 2
[NO]
[NO] * 3
END
[YES]
WAIT
*6
(Stabilization time)
*5
[YES] or [NO]
Measurement mode
F0902.ai
*1: See "Figure 6.1 Modes of Operation Level" on page 6-8.
*2: Press the [NO] key to return to the parameter for sample gas.
*3: Using the [NO] key, show [V_ZERO], [V_SPAN] and [END] in succession.
*4: Using the [YES] key, open the zero-gas or span-gas value to read the measured value
at the point.
*5: Press the [YES] key to go into the measurement mode state after a given stabilization
time elapses; press the [NO] key to go into that state immediately.
*6: This step is skipped if no valve was opened.
*7: The display shows the same data item to be measured as the one in the measurement mode.
Figure 9.1
Valve Operation Mode
IM 11T03E01-01E
<9. CALORIE / CALIBRATION PROCEDURE>
9.2.2
9-3
Setting Calibration Data
To set calibration data, press the [*] key in the measurement mode, (when password is selected,
enter the password (XXX), ) and select the parameter [*CAL.DT]. See Figure 6.4 for the process
at the setting level.
(a) Zero-point calorific value [*Z_CAL]
Configurable range : 000.000 to 133.000 (MJ/m3)
(b) Zero-point density [*Z_DNS]
Configurable range :
0.0000 to 3.5000 (kBTU/ft3)
0.0000 to 6.0000 (kg/m3)
0.00000 to 0.40000 (lb/ft3)
(c) Span-point calorific value [*S_CAL]
Configurable range : 000.000 to 133.000 (MJ/m3)
(d) Span-point density [*S_DNS]
Configurable range :
0.0000 to 3.5000 (kBTU/ft3)
0.0000 to 6.0000 (kg/m3)
0.00000 to 0.40000 (lb/ft3)
(e) Output hold value during calibration [*C_HLD]
• Disable:
0
• Enable:
1
• Enable (preset value): 2
(f) Setting of preset value [*PR.SET]
Configurable rang :
-10.0 to 110.0
(Applies only if item (e) above is set to “2”.)
IM 11T03E01-01E
<9. CALORIE / CALIBRATION PROCEDURE>
*1
*RANGE
[NO]
*CAL.DT
[NO]
9-4
*ALARM
[YES]
XXX.XXX
*Z_CAL * 2
(Setting of Zero-point calorific value)
[ENT]
X.XXXX
*Z_DNS * 2
(Setting of Zero-point density)
[ENT]
XXX.XXX
*S_CAL * 2
(Setting of Span-point calorific value)
[ENT]
X.XXXX
*S_DNS * 2
(Setting of Span-point density)
[ENT]
X
*C_HLD
*2
(Setting of Output hold value during calibration)
[ENT]
XXX.X
*PR.SET
*3
*2
(Setting of preset value)
[ENT]
F0903.ai
*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*3: This step is skipped if the parameter [*C_HLD] is set to a value other than "2".
Figure 9.2
Calibration Parameter Setting Mode
IM 11T03E01-01E
<9. CALORIE / CALIBRATION PROCEDURE>
9-5
9.3 Calibration
9.3.1
Semi-automatic Calibration
In this mode of calibration, you can carry out one-touch calibration. To use this feature, you must
observe the following instructions.
• The analyzer needs solenoid valves for controlling the calibration gases (zero and span
gases). Connect the solenoid valves with utmost care to ensure that there is no gas leakage
in the system.
• Install the calibration gas cylinders in a location close to the GD40 detector. If the detector
needs to be installed in an explosion-hazardous area, the solenoid valves being used must
also be protected against explosion.
• The contacts for controlling solenoid valves are built in the GD402T, GD402V, GD402R
converter. It is the user's responsibility to supply power to drive the contacts. You must install
a switch and a fuse along the power line. Note that the contacts are rated at 250 V AC/3 A or
30 V DC/3 A.
*1
DISP
[NO]
[NO]
SEM.CAL
MAN.CAL
[YES]
[NO]
START
[YES]
XX.XXXX
ZERO
XX.XXXX
SPAN
XX.XXXX
WAIT
* 34
*5
*
Zero-point density display
*3
*5
Span-point density display
*2
*3
Measurement mode
Stabilization time
*2
F0904.ai
*1: See" Figure 6.1 Modes of Operation Level" on page 6-8.
*2: If a calibration data error [ALM.10] occurs, the display alternates between the [WAIT] and [ALM.10] indications.
The [ALM.10] indication still remains on display even after the mode move to the measurement mode.
This error massage does not disappear until normally ending in re-calibration or resetting the analyzer.
*3: The values are operated and the calibration data are read at the time setpoint (calibration time) which is defined
in the mode for setting automatics/semi-automatic calibration data in CODE 15.
The display always shows the corrected density.
*4: During automatic calibration, the display shows the same data items as those of the [ZERO] and subsequent
indications.
*5: When pressing the [MODE] key during [ZERO] or [SPAN] status, the [WAIT] appears on the display.
When pressing the [MODE] key once again, the analyzer returns to the measurement mode.
(This operation means you have cancelled automatic/semi-automatic calibration.)
Figure 9.3
Semi-automatic Calibration Mode
IM 11T03E01-01E
<9. CALORIE / CALIBRATION PROCEDURE>
9.3.2
9-6
Manual Calibration
In this mode of calibration, you visually make sure the reading has settled, and then confirm the
reading manually (by pressing the [ENT] key). There are two ways to switch between the zero,
span and sample gases: by using manually-operated valves or by using solenoid valves. When
switching between the gases using solenoid valves, follow the instructions on valve installation in
Subsection 9.3.1.
*1
SEM.CAL
[NO]
MAN.CAL
[NO]
[YES]
VALVE
[NO]
ZERO
[YES]
XXX.XXX
Z_CAL
[ENT]
Zero-point calorific value display
*3
X.XXXX
Z_DNS
[ENT]
Span-point density display
*3
X.XXXX
CAL.SET
*6
[NO]
[ENT]
*4
SPAN
[YES]
XXX.XXX
S_CAL
[ENT]
Span-point calorific value display
*3
X.XXXX
S_DNS
[ENT]
Span-point density display
*3
X.XXXX
CAL.SET
[ENT]
*4
*2
ALM.10
X.XXXX
WAIT
[YES] or [NO]
Measurement mode
*5
F0905.ai
*1: See "Figure 6.1 Modes of Operational Level" on page 6-8.
*2: If a calibration data error occurs, the display shows [ALM.10].
Pressing the [YES] or [NO] key returns the display to the [MAN.CAL].
*3: Set a desired value by using the [ ], [ ] and [ENT] keys.
*4: When the reading settles, press the [ENT] key to confirm your calibration.
This parameter does not accept input with the [ ] and [ ] keys.
If you press the [MODE] key (for ESCAPE) after this parameter appears, the display shows [WAIT].
The display always shows the corrected density.
*5: Press the [YES] key to go into the measurement mode status after a given stabilization time elapses: Press the [NO]
key to go into that state immediately.
*6: In case of pressing the [NO] key after skipping zero calibration (by pressing [NO] key in the [ZERO] status),
the analyzer returns to [MAN.CAL].
In case of pressing the [NO] key after finishing zero calibration, the analyzer goes to [WAIT] status.
Figure 9.4
Manual Calibration Mode
IM 11T03E01-01E
9.3.3
<9. CALORIE / CALIBRATION PROCEDURE>
9-7
Automatic Calibration
This mode of calibration takes effect when you select “enable” for automatic calibration in CODE
13 - [*AUTO.C] - at the service level. Even during automatic calibration, you can interrupt the
system with the semi-automatic or manual mode of calibration. One cycle of automatic calibration
is skipped, however, if the time of automatic calibration arrives when semi-automatic calibration is
in progress.
Measurement mode
X.XXXX
ZERO
X.XXXX
SPAN
*1
*1
X.XXXX
WAIT
*2
Measurement mode
F0906.ai
*1 : Pressing the [MODE] key, change it to display [WAIT] indication.
*2 : Pressing the [MODE] key, once again returns it measuring mode.
Figure 9.5
Automatic Calibration Mode
IM 11T03E01-01E
Blank Page
<10. HYDROGEN PURITY / CALIBRATION PROCEDURE>
10-1
10. HYDROGEN PURITY /CALIBRATION
PROCEDURE
This chapter describes how to calibrate of Hydrogen purity meter.
If select Density meter, see Chapter 8.
If select Calorie meter, see Chapter 9.
Zero and span calibration can be carried out using standard gases (zero and span gases). The
analyzer uses hydrogen (100%) as the zero gas and carbon dioxide (100%) as the span gas.
The analyzer is designed so that calibration is done manually. The basic calibration procedures
and operations are as follows.
10.1
Basic Calibration Procedure
Operation level
Manual calibration mode
See the Process in Figure 7.2.
See the Process in Figure 10.2.
F1001.ai
10.2
Preparation for Calibration
Prepare for calibration during the initial operation of the analyzer by entering manual calibration
at the operation level.
*1
*RANGE
[NO]
[NO]
*CAL.DT
*ALARM
[NO]
*SERVC
[NO]
[YES]
X
*C_HLD
*2
(Setting of output hold value
during calibration)
[ENT]
XXX.X
*PR.SET
*2
*3
(Setting of preset value)
[ENT]
F1002.ai
*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*3: This step is skipped if the parameter [*C_HLD] is set to a value other than "2".
Figure 10.1
Calibration Parameter Setting Mode
IM 11T03E01-01E
10-2
<10. HYDROGEN PURITY / CALIBRATION PROCEDURE>
10.3 Calibration
In the manual calibration mode, visually check that the reading has settled, and then confirm the
reading manually (by pressing the [ENT] key).
*1
DISP
[NO]
[NO]
MAN.CAL
S_GAS
[YES]
ZERO
* 5 [NO]
Close the sample gas valve.
[YES]
0.0899
H2
*3
Zero-point density (H 2 gas) is displayed.
[ENT]
X.XXXX
CAL.SET
*4
[ENT]
*
5
[NO]
Open the zero gas valve.
Press [ENT] key after the value is stabilized.
Close the zero gas valve.
SPAN
[YES]
1.9771
CO2
*3
[ENT]
X.XXXX
CAL.SET
*4
[ENT]
Span-point density (CO 2 gas) is displayed.
Open the span gas valve.
Press [ENT] key after the value is stabilized.
Close the span gas valve.
Open the sample gas valve.
*2
ALM.10
Measurement mode
*1: See "Figure 7.2 Modes of Operational Level" on page 7-7.
*2: If a calibration data error occurs, the display shows ALM.10 .
Pressing the [YES] or [NO] key returns this status to the [MAN.CAL].
*3: The displayed value is fixed. It cannot be set.
*4: When the reading settles, press the [ENT] key to confirm your calibration.
This parameter does not accept input with the [ ] and [ ] keys.
The display always shows the corrected density.
*5: Pressing the [NO] key for both the [ZERO] and [SPAN] indications returns
this status to the [MAN.CAL].
Figure 10.2
F1003.ai
Manual Calibration Mode
IM 11T03E01-01E
11-1
<11. INSPECTION AND MAINTENANCE>
11. INSPECTION AND MAINTENANCE
To ensure the GD402 analyzer remains highly accurate with excellent operating conditions,
routine inspection and maintenance is essential. Refer to the check items in this chapter to
determine which items to apply in your inspection and maintenance, and check them periodically.
11.1
11.1.1
Routine Inspection and Maintenance
Checking Readings and Calibrating the Analyzer
Measure the standard gases every two to three months to check the output signal (Note: the
interval varies depending on the operating conditions). If the output signal has errors, carry out
zero and span calibration. For details on calibration, see Chapter 8, 9 or 10.
11.1.2
Checking the Flowrate of Sample Gas
The sample gas should be supplied to the detector at 600 mL/min ±10%. Periodically make sure
the given flowrate is kept constant. When checking the flowrate, also make sure the piping is
normal and there is no leakage of calibration gases.
11.1.3
Periodic Replacement of the Detector's O-rings
The O-rings are made from NBR, a material that is not very susceptible to plastic deformation.
Their sealing capabilities therefore are not readily impaired. Any deterioration in these O-rings
however can result in not only gas leakage but reduced resistance to mechanical vibration,
which in turn may lead to reading errors. For this reason, it is recommended that the O-rings be
replaced at a fixed interval (every two to three years). Consult Yokogawa for the replacement of
O-rings.
IM 11T03E01-01E
11.1.4
11-2
<11. INSPECTION AND MAINTENANCE>
Replacing the Fuse
(1) For safety reasons, stop the power supply to the converter by switching the external
breaker.
(2) Remove the fuse from the fuse holder. Turn the holder cap counterclockwise by 90 degrees
using a flat-blade screwdriver that fits the holder cap. Then, pull out the fuse together with
the cap.
(3) Check that the rating of the new fuse is correct, and put it in the fuse cap; insert the cap (and
fuse) in the holder; then while pressing down, turn the cap clockwise by 90 degrees using
the flat-blade screwdriver.
(4) If the new fuse breaks too soon, the circuit may be abnormal. Contact Yokogawa in such a
case.
Fuse Holder
General purpose type
Fuse Holder
Explosion-proof type
Figure 11.1
F1101.ai
Position of the Fuse Holder
11.1.5 Cleaning
Use a clean, soft cloth when cleaning for inspection or maintenance. You can use a neutral
detergent on the transparent window of the non-explosion-protected converter (weatheringresistant PVC sheet), when extremely dirty. But do not use any organic solvent.
If the device is so dirty and/or badly scratched that it affects key operation or reduces visibility,
replace the case cover (part No.: K9313DW).
IM 11T03E01-01E
11-3
<11. INSPECTION AND MAINTENANCE>
11.2
Inspection In Case of Failure
11.2.1
Inspecting the Analyzer in an Alarm Status
The GD402 analyzer outputs an ALARM signal through terminals 16 and 17 if a failure is
detected by the self-diagnostic function during measurement. In addition, the lamp
on the operation panel comes on and the message field shows an alarm number. The analyzer
outputs no contact signal, however, if the alarm number is [ALM.09]. If a failure (alarm) is
detected, take the corrective action given in Table 11.1.
Table 11.1
Message
Shown
Corrective Actions when an Alarm Status Is Detected
Failure Type
Diagnostic Conditions
Corrective
Actions
Message via
Communication
ALM.01
High/low limit alarm
against physical
density
Physical density ≤ Low-limit-ofChange the high/low
physical-density alarm set-point limits.
Physical density ≥ High-limit-ofphysical-density
OUT OF DENSITY
ALM.02
High/low limit alarm
against compensated
density
Compensated density ≤ Low-limit-of- Change the high/low
compensated-density alarm set-point limits.
Compensated density ≥ High-limit-ofcompensated-density alarm set-point
OUT OF COMP
DENSITY
ALM.03
High/low limit alarm
Specific gravity ≤ Low-limit-of-specific Change the high/low
against specific gravity
gravity alarm set-point limits.
Specific gravity ≥ High-limit-of-specific
gravity alarm set-point
OUT OF SP GR
ALM.04
High/low limit alarm
against calorific value
Calorific value ≤ Low-limit-of-calorific- Change the high/low
value alarm set-point limits.
Calorific value ≥ High-limit-of-calorificvalue alarm set-point
OUT OF CALORY
ALM.05
High/low limit alarm
against molecular
weight
Molecular weight ≤ Low-limit-ofChange the high/low
molecular weight alarm set-point limits.
Molecular weight ≥ High-limit-ofmolecular weight alarm set-point
OUT OF
MOLECULE
ALM.06
High/low limit alarm
against concentration
Concentration ≤ Low-limit-ofChange the high/low
concentration alarm set-point limits.
Concentration ≥ High-limit-ofconcentration alarm set-point
OUT OF CONC
ALM.07
Error in pressure input
range
Zero point -3% ≥ pressure
Span point +5% ≤ pressure
Pressure ≤ 0.1 kPa
Check the sample-gas OUT OF GAS
pressure and the
PRESS
pressure range.
ALM.08
Abnormal sample gas
temperature
-25 °C ≤ sample gas temperature
≤ 80°C
Use the sample gas
within the tolerance
limits.
OUT OF GAS
TEMP
ALM.09
Battery failure
Contact Yokogawa
service personnel.
LOW BATTERY
ALM.10
Error in calibration
(zero and span points)
Redo calibration.
ILLEGAL CALIB
-0.3000 ≤ zero calibration value
≤ 0.3000
0.50000 ≤ span calibration value
≤ 1.5000
IM 11T03E01-01E
11-4
<11. INSPECTION AND MAINTENANCE>
11.2.2
Inspecting the Analyzer in a FAIL Status
IN GD402, the state of terminals 18 and 19 is changed from close to open if a FAIL status is
detected by the self-diagnostic function during measurement. In addition, the lamp on
the operation panel comes on and the message field shows an error number. If a FAIL status is
detected, take the corrective action given in Table 11.2.
Table 11.2
Message
Shown
Corrective Actions when a FAIL Status Is Detected
Failure Type
Diagnostic Conditions
Corrective Action
Message via
Communication
Err.01
Sensor oscillation
shutdown
Failure to detect
oscillation frequency from the
detector
Reset the power
supply.
Contact Yokogawa
service personnel.
FAULT
SENSOR
Err.02
Error in sensor’s
oscillation
frequency
Frequency F2 : out of
range from 1000 to 10000Hz
Frequency F4 : out of
range from 4000 to 10000Hz
Reset the power
supply.
Contact Yokogawa
service personnel.
OVER
FREQUENCY
Err.03
Error in sensor’s
temperature
Sensor’s failure to detect
temperature
Contact Yokogawa
service personnel.
FAULT TEMP
Err.04
A/D converter
failure
Failure in A/D converter
Contact Yokogawa
service personnel.
FAULT A/D
Err.05
Memory failure
Erroneous data in
EEPROM/EPROM/RAM
Contact Yokogawa
service personnel.
FAULT
MEMORY
IM 11T03E01-01E
Customer
Maintenance
Parts List
6
5
1
GD40G, GD40T, GD40V, GD40R
Gas Density Meter
Detector
3
2
4
7, 8
9
Item
1
2
3
4
5
6
7
8
9
Part No. Qty
F9301DY 1
F9301DQ 1
G9303AM 1
G9303LK 1
G9339AA 2
Description
Cover
Cover
O-ring
O-ring
Bracket Assembly
Y9408ZU
F9340AL
Y9401WL
G9601AN
Bolt
Screw
Toothed Lock Washer
Cable Gland Assembly (only GD40R)
2
1
1
1
All right reserved, Copyright © Copyright 1998,
Yokogawa Electric Corporation.
Subject to change without notice
CMPL 11T3E1-01E
1st Edition : Apr. 1998 (YK)
5th Edition : Jun. 2015 (YK)
Customer
Maintenance
Parts List
GD402G
Gas Density Meter
Non-Explosion-proof Converter
1
6
2
5
4
3
2
3
4
Part No. Qty
1
A1109EF
1
A1111EF
6
L9811FV
K9334CN 6
K9313DW 1
Description
Fuse 1.0 A (For 100 to 240 V AC)
Fuse 2.0 A (For 24 V DC)
Cable Gland
Insert
Cover Assy
5
6
Y9101XA
Y9420LU
O-Ring
Screw
Item
1
4
4
All right reserved, Copyright © Copyright 1998,
Yokogawa Electric Corporation.
Subject to change without notice
CMPL 11T3E1-02E
1st Edition : Apr. 1998 (YK)
3rd Edition : Apr. 2012 (YK)
Customer
Maintenance
Parts List
3
GD402T, GD402V, GD402R
Gas Density Meter
Explosion-proof Converter
2
7
1
4
5
6
Item
1
2
3
4
Part No. Qty
A1109EF
1
A1111EF
1
L9817MH 1
K9292XL 1
L9800GT 1
Description
Fuse 1.0 A (For 100 to 240 V AC)
Fuse 2.0 A (For 24 V DC)
O-Ring
Screw
Screw
5
6
7
K9214GZ
L9811LL
K9214BF
Plug
Cable Gland G3/4 (PF3/4) (only GD402R)
Cover
3
3
1
All right reserved, Copyright © Copyright 1998,
Yokogawa Electric Corporation.
Subject to change without notice
CMPL 11T3E1-03E
1st Edition : Apr. 1998 (YK)
3rd Edition : Apr. 2012 (YK)
i
Revision Information
Title
: Model GD402 Gas Density Meter
Manual No. : IM 11T03E01-01E
Edition Date
Remark(s)
11th June 2017
Added RoHS compliance statement.
10th
Jan. 2017
EJA310 to EJX310A (Pages 1-2, 1-3, 1-6, 1-7, 1-8, 1-12, 2-11, 2-20)
Addition of Applicable standards (Pages 1-1 and 1-2)
Revised over all
9th
Feb. 2016
Addition the related documents, etc.
IM number change (IM 11T3E1-01E to IM 11T03E01-01E)
8th
Jun. 2015
Style change
Page i
Pages 1-3, 1-6 Change of fuse type
Pages 1-7, 1-8 Addition of “Style”
Pages 1-7, 1-14, 2-14, 2-16
Deletion of “Surge Protector”
Correction
Pages i, 1-4, 1-7, 1-8, 1-12, 2-11, 2-12, 2-14, 2-16, 5-5, 5-24, 6-5, 6-24, 10-2
CMPL 11T3E1-01E
5th edition
7th
June 2012
Revised and Corrected over all
6th
Sep. 2008
All over revised
5th
Jul. 2006
Revised Section
After-sales Warranty: Added description.
Added Section 1.1.1.7, “Safety and EMC compliance”
Deleted Section 1.4, “Safety Standard”
Section 2.1.2, “GD40T (FM Explosion-proof with Intrinsically Safe Approval)”:
Correction error in Item 1.
Section 2.4.1, Replaced the entire page.
Replaced entire Section 2.4.2, “Cable Wired to Power Supply”
Section 5.2.2 “Example of Entries for Concentration Measurement”:
Correction error of *Z_CN.D and *S_CN.D.
CMPL 11T3E1-01E:
Deleted Item 7, moved up the following Items, and changed part numbers of
Items 5 and 9.
4th
May 2004
1st
Sep. 1998
Newly published
n If you want have more information about Yokogawa products, you can visit
Yokogawa’s home page at the following website.
http://www.yokogawa.com/an/
n Published by Yokogawa Electric Corporation
2-9-32 Nakacho, Musashino-shi, Tokyo 180-8750, JAPAN
n Printed by
KOHOKU PUBLISHING & PRINTING INC.
IM 11T03E01-01E
Blank Page