Extractive CEMS Presentation
Industry Manual Repository
Join the AnalyzeDetectNetwork and Read This Manual and Hundreds of Others Like It! It's Free!
INFRARED BAND CENTERS
OF SOME COMMON GASES
Gas
NO
Band Center
( m)
5.0-5.5
Wave Number
(cm-1)
1800-2000
NO2
5.5-20
500-1800
SO2
8-14
700-1250
H2O
3.1, 5-5.5,
7.1-10
2.7, 5.2, 8-12
10.5
1000-1400
1800-3200
850-3700
950
CO2
NH3
GENERAL CATEGORIES OF CEMS
OPACITY
SINGLEPASS
SYSTEMS
DOUBLEPASS
SYSTEMS
GASEOUS EMISSION MONITORS
EXTRACTIVE
SYSTEMS
DILUTION
IN-SITU
SYSTEMS
REMOTE
SYSTEMS
IN-STACK
CROSS-STACK
BASIS OF CEM ANALYSIS
• Extractive
Dry, Wet, or Dilution
• In-situ
Wet basis
• Dilution
Wet
ANALYSIS CONDITIONS
• Extractive
• In-Situ
• Dilution
• Analysis performed at
controlled conditions
(constant temperature
and pressure)
• Analysis performed at
stack conditions
(compensation for, or
approximation of,
temperature and
pressure)
• Analysis performed at
controlled conditions
CEM CALIBRATION
METHODS
• Extractive
• Compressed gases
• Internal electronic
• In-Situ
• Gas cell and
compressed gases for
point/in-situ
• Internal electronic
• Dilution
• Compressed gases
• Internal electronic
ADVANTAGE OF EXTRACTIVE
CEMS
• Allows widest selection of analyzer
technologies
• Can analyze at ambient conditions for
which more reference data are available
• Can combine more than one analyzer
(e.g., GC and FID)
• Can remove interfering substances before
measurement
• Gas is measured on a dry basis
ADVANTAGE OF EXTRACTIVE
CEMS
• Analyzers can be installed in an
accessible, clean environment
• Multi-probe capability for representative
sample
• Perform multiple point sampling via timesharing
• Expands easily for additional sampling
points
• Is relative accessible for service and
maintenance
DISADVANTAGE OF
EXTRACTIVE CEMS
• Sample transport and conditioning system
is expensive to install and operate and has
high power requirements
• Sample transport and conditioning system
has potential for pluggage, leaks and
condensation problems (both water and
acid)
• Gas conditioning is often required
• May inadvertently remove substances of
interest from sample gas
DISADVANTAGE OF
EXTRACTIVE CEMS
• Condensed water and/or filter residues
may need to be analyzed
• Has more components and a more
complicated design
• Has time-lag during analysis
ADVANTAGES OF IN-SITU
PATH CEMS
•
•
•
•
•
•
•
Fast response time
No calibration gas needed
No sample transport or conditioning
Simple.less expensive installation
More representative gas measurement
Less equipment to buy and maintain
Has fewer components
DISADVANTAGES OF IN-SITU
PATH CEMS
• Potential interference by particulate or
droplets
• Gas measured on wet basis,moisture
content must be assumed
• Sometimes cannot locate equipment
downstream of sorbent injection or spray
dryer systems
• Analyzer is exposed to harsh operating
conditions and vibrations
DISADVANTAGES OF IN-SITU
PATH CEMS
• Limits choice of analyzers
• Cannot perform a dynamic calibration
when process is operating
• Is relatively inaccessible for service and
maintenance
• Does not allow for expansion
ADVANTAGES OF IN-SITU
POINT CEMS
•
•
•
•
•
Fast response time
No sample transport or conditioning
Simple, less expensive installation
Less equipment to buy and maintain
Has few components
DISADVANTAGES OF IN-SITU
POINT CEMS
• Representative sample difficult to obtain in some
situations
• Gas measured on wet basis
• Vibration sensitive
• Access for service and maintenance can be
difficult
• Limits choice of analyzer
• Does not allow for expansion
• Operates in a potentially harsh environment
ADVANTAGES OF DILUTION
CEMS
• Allows wide selection of analyzers
• Can analyze at ambient conditions for which
more reference data is available
• One dilution system can serve several analyzers
• Analyzers can be installed in an accessible,
clean environment
• Multi-probe capability for representative sample
• Heated sample lines and moisture removal
system not necessary
DISADVANTAGES OF
DILUTION CEMS
• Measurement accuracy and data precision
problems may occur with highly diluted samples
• Dilution system may not work on high moisture
flue gas
• Gas is measured on a wet basis; this may not
be a problem if CO2 is used as the dilution gas
• Requires additional calibration for the dilution
system
EXTRACTIVE SYSTEM
COMPONENTS AND ACCESS.
•
•
•
•
•
•
Sample Probe: Const./Composition
Probe Blowback: Design and Frequency
Sample Line: Comp./Length/Diameter
Valves/Fittings: Const./Composition
Pressure/Vacuum Meters: Quality
Moisture Conditioning System:
Refrigerated, Dilution, Capacity, Design,
Construction
EXTRACTIVE SYSTEM
COMPONENTS AND ACCESS.
• Filters: Coarse/Fine and Quality
• Pumps: Capacity, type, Quality
• Cabinets or Shelters: Location,
Temperature Stability
• System Controller: Microprocessor To
Sequence/Control Automatic Functions
• Electrical Support: Fuses, Circuit
Breakers
• Calibration Gases: Location, Injection
Point, tubing Regulations, Gas
Parameter
COMPARISON
GFC
NDIR
• Energy Source
Nichrome Coil
Nichrome Coil
• Spectrometer
NDIR
NDIR
• Reference Cell
100% CO
0% CO
• Sensivity
Superior due to more
Inferior due to Utilization
energy reaching detector
of a Single Absorption
and ability to utilize Multi- peak and Interference from
Pass Optics
CO2 & H2O
• Drift
• Specificity
Reduced due to
Higher due to two cell
cancellation of light
Intensity variations
design
Interferents absorb in CO2 & H2O interferes
measure & reference due to similar absorption
beams equally
characteristics as CO
MODEL 48C-CO Analyzer and
41CHL CO2 Analyzer
• Based on EPA Reference
Method (RFCA-0981-054)
• Gas Filter Correlation
Principal
• Microprocessor control
• Remote Diagnostics
• No consumables
GENERIC INFRARED CELL
Detector
Sample
Filter
Out
Sample Cell
In
Reference Filter
Light Source
MODEL 48C Flow
Selected D*etector & Pre-Amp
Span
CO Reactor
Multi-Pass Sample
Thermostated Cell
Gold Optics
Flow
Sensor
Sample
Pressure
Sensor
Band-Pass Filter
Permeation
Dryer
Chopper
Correlation Wheel
Purge Housing
IR
Source
Chopper
Motor
Sample Pump
Thermo Model 41C CO2
Analyzer
MODEL 43C- SO2 ANALYZER
• Microprocessor control
• SO2 Specific
• Reflective U.V. filtering
• Hermetically sealed U.V.
lamp
• No consumables
MODEL 43C-SO2 Analyzer
SO2 + h
Ia
1
SO2*
Kf
SO2*
SO2 + h
2
SAMPLE IN
Ia = Io[I-e-{ax(SO2)}]
If
Ioax(SO2) or
K (SO2)
SAMPLE OUT
PHOTO
MULTIPLIER
TUBE
FLASHER
SUPPLY
ELECTRONICS
MODEL 43C- FLOW SCHEME
Power Supply
Signal Processing
Band-pass
Filter
PMT Detector
Photo-detector (feedback)
U.V. Lamp
Sample
Pressure
Transducer
HC Kicker
Span
Zero
Capillary
Flow
Transducer
Span/Zero
Valve
Option
Sample Pump
Exhaust
Thermo Model 43C SO2 Analyzer
15
HOW IS NO MEASURED?
Chemiluminescence Technique
NO + O3
NO2* + O2
NO2*
NO2 + h
Intensity of emitted light is proportional to
NO concentration
Chemiluminescent
Reduction of NO2 to NO
3 NO2 + Mo
3 NO + MoO3
A molybdenum catalyst, heated to ~325oC, is
used to Convert NO2 to NO
LS 420 Insitu O2
Oxygen Sensor
Oxygen molecules tend to flow from the reference side with the highest
concentration of O2 to the sample side with the lowest O2 concentration.
Within the solid electrolyte the flow of oxygen generates a voltage which
is related to the concentration of O2 in the sample by the Nernst equation.
%O2=20.95*10 exp(-E/C)
Zirconium Oxide
Electrolyte
Gas Flow
o
850 C
-
O2
-
O2
E
21% O2
O2
+
O2
O2
O2
O2
O
O2
O2
O=
O2
O2
Electrocatalytic
Platinum Surface
=
O
O2
O2
=
O2
<21%
O2
O2
LS420 Oxygen Analyzer
• Automatic Calibration
– Requires Auto Cal Unit
– Calibrates in synchronism with
LS710/SM8175
• Same proven in situ features
– Low maintenance and high reliability
– Calibration under process temp and press
– Excellent accuracy and specificity
– Easily integrated into CEMS package
DIFFERENTIAL
GAS ANALYZER CEMS BIASES
BIASES
Problem PRESSURE
Corrective
Action
•Other Gases Interfere
With Measurement Gas
•Features of
Instrumentation
Produces Biases
•Change Analytical
Technique
•Scrub Out Interfering
Gas
•Math Correction
•Choose Analyzers
Wisely
•Design QA/QC
Program to Address
Design Features
DIFFERENTIAL
GAS ANALYZER CEMS BIASES
BIASES
Problem PRESSURE
Corrective
Action
•Exposing Analyzers to
Extreme Temperatures
and Changing
Barometric Pressure
•Plume Downwash or
Flue Gas Exhaust Into
CEM Shelter Causes
Systematic Error
•Temperature Stabilize
the Analyzer and
Mathematically
Compensate for
Pressure Effects
•Shelter or Otherwise
Protect System
•Filter Ambient Air
DIFFERENTIAL
GAS ANALYZER CEMS BIASES
Problem PRESSURE
Corrective
Action
BIASES
•Incorrect Gas Values
•Replace or Recertify
Used During Calibration Gas Standard
•Inadequate/Inconsistent •Establish Procedures
Response Time
That Ensure Adequate
Time For Monitor
Response
•System Calibration
•Perform Probe and
Obscuring Local Biases Local Analyzer
Calibration Checks In
Addition System Check
SOURCES OF BIASES IN
DILUTION PROBES
• After wet scrubbers, aerosols enter the probe
and changes dynamics of dilution
• Dilution probes are affected by changes in Ts,
Ps, and Ms which changes dilution ratio
• Calibration of dilution probes changes from
calibration “gas blends” vs. “single” gases (i.e.,
CO2 in the blend changes Ms of the gas blend!)
• Contaminated source “dilution air” effects
dilution ratio
ADVANTAGES OF USING
DILUTION PROBES
• Meets 40CFR75 emission requirements of
reporting on a “wet” bases
Pmr = CwAsvw
• Sampling rate (~20-50 mL/min) much better than
conventional extractive systems (~2-5 L/min.)
• Reduces moisture of the sample gas, this not
requiring “heated” sample lines to analyzer,
therefore lower maintenance
• Use ambient monitors which meet design and
performance criteria set by EPA